Wavetek 2015 Manual de usuario

Categoría
Multimetros
Tipo
Manual de usuario
®
TM
2015.qxd 7/21/00 5:40 PM Page 1
WARRANTY
The 2015
Digital Multimeter is warranted against any defects of material or
workmanship within a period of one (1) year following the date of purchase of the
multimeter by the original purchaser or original user.
Any multimeter claimed to be defective during the warranty period should be returned
with proof of purchase to an authorized Wavetek Meterman Service Center or to the local
Wavetek Meterman dealer or distributor where your multimeter was purchased. See
maintenance section for details.
Any implied warranties arising out of the sale of a W
avetek Meterman multimeter,
including but not limited to implied warranties of merchantability and fitness for a
particular purpose, are limited in duration to the above stated one (1) year period.
Wavetek Meterman shall not be liable for loss of use of the multimeter or other
incidental or consequential damages, expenses, or economical loss or for any claim or
claims for such damage, expenses or economical loss.
Some states do not allow limitations on how long implied warranties last or the
exclusion or limitation of incidental or consequential damages, so the above limitations
or exclusions may not apply to you.
This warranty gives you specific legal rights, and you may also have other rights
which vary from state to state.
D GEWÄHRLEISTUNG
Die Digitale Multimeter Modelle 2015 ist ab Kaufdatum für ein (1) Jahr
gegen Material- und Herstellungsfehler gewährleistet. Siehe Kapitel "Reparatur"
für Einzelheiten.
Implizierte Schadeforderungen sind auch auf ein Jahr beschränkt. W
avetek
Meterman ist nicht ansprechbar für Gebrauchsverluß oder Folgeschäden,
Ausgaben, Gewinnverluß, usw.
E GARANTIA
Este Multímetro Digital Modelo 2015 está garantizado contra cualquier defecto de
material o de mano de obra durante un periodo de un (1) año contado a partir de la
fecha de adquisición. En la sección de "Reparación" se explican los detalles relativos a
reparaciones en garantía.
Cualquier otra garantía implícita está también limitada al periodo citado de un (1)
año. Wavetek Meterman no se hará responsable de pérdidas de uso del multímetro, ni
de ningún otro daño accidental o consecuencial, gastos o pérdidas económicas, en
ninguna reclamación a que pudiera haber lugar por dichos daños, gastos o pérdidas
económicas.
F GARANTIE
Le multimètre digitaux, Modèle 2015 est garanti pour un (1) an à partir de la date
d'achat contre les défauts de matériaux et de fabrication. Voir chapitre "Réparation" pour
plus de détails.
T
oute garantie impliquée est également limitée à un an. Wavetek Meterman ne peut
être tenu responsable pour perte d'utilisation ou autres préjudices indirects, frais, perte
de bénéfice, etc.
2015.qxd 7/21/00 5:40 PM Page 2
CONTENTS
D • Inhalt
E • Contenido
F • Contenu
EXPLANATION OF SYMBOLS
D • Erklärung der Symbole = E • Significado de los símbolos = F • Explication des Symboles
Safety Information ............................................... 2
Instrument Familiarization.................................... 5
Measurement Procedures ................................... 8
Specifications ................................................... 28
Troubleshooting and Maintenance .................... 37
Sicherheitsinformationen .................................... 2
Vorstellung des Gerätes....................................... 5
Meßprozeduren ................................................... 9
Spezifikationen .................................................. 30
Fehlersuche und Unterhalt ................................ 37
Información de seguridad ................................... 3
Familiarización con el instrumento ...................... 5
Procedimientos de medida .................................. 9
Especificaciones ............................................... 32
Localización de avería y mantenimiento............. 37
Informations de Sécurité ..................................... 3
Présentation de l’Appareil ................................... 5
Procédures de Mesure ...................................... 10
Spécifications ................................................... 34
Dépannage et Maintenance................................ 37
Attention! Refer to Operating Instructions •D•
Achtung! Bitte Anleitung
lesen
•E• ¡
Atención! Consulte las Instrucciones de Uso
•F•
Attention!
Consultez le manuel.
Ground connection •D•
Erdanschluß
•E•
Conexión a tierra
•F•
Connection de terre.
Alternating current •D•
Wechselstrom
•E•
Corriente alterna
•F•
Courant alternatif.
Direct current •D•
Gleichstrom
•E•
Corriente continua
•F•
Courant continu.
Dangerous voltage may be present at terminals •D•
Eine gefährliche
Spannung kann an den Eingängen anliegen
•E•
Puede haber tensión
peligrosa en los terminales
•F•
Une tension dangereuse peut être
présente aux entrées.
This instrument has double insulation •D•
Dieses Gerät ist doppelt
geisoliert
•E•
Este instrumento tiene doble aislamiento
•F•
Cet
appareil est prévu d’une double isolation.
– 1 –
2015.qxd 7/21/00 5:40 PM Page 3
WARNINGS AND PRECAUTIONS
n This instrument is EN61010-1 certified for Installation Category III. It
is recommended for use in distribution level and fixed installations, as
well as lesser installations, and not for primary supply lines, overhead
lines and cable systems. n All inputs are protected against continuous
overload conditions up to the limits of each function's stated input
protection (see Specifications). Never exceed these limits
nor the ratings
marked on the instrument itself. n Voltages applied to COM terminal
may be present at all other terminals n Exercise extreme caution when:
measuring voltage >20V, current >10mA, AC power line with inductive
loads, AC power line during electrical storms. High voltages can be lethal
and high voltage transients may occur at any time. n Operator injury or
damage to the multimeter may occur during current measurements if the
fuse blows in a circuit with open circuit voltage >600V (250 V in mA
input). n Always inspect your DMM, test leads and accessories for
sign of damage or abnormality before every use. If abnormal condition
exist (broken or damaged test leads, cracked case, display not reading,
etc.), do not use. n When testing for voltage or current, make sure these
ranges function correctly. Take a reading of a known voltage or current
first. n Never ground yourself when taking measurements. Do not
touch exposed metal pipes, outlets, fixtures, etc., which might be at
ground potential. Keep your body isolated from ground and never touch
exposed wiring, connections, test probe tips, or any live circuit
conductors. n Always measure current in series with the load – NEVER
connect the multimeter ACROSS a voltage source. Check fuse first. n
Never replace a fuse with one of a different rating. n Do not operate
instrument in an explosive atmosphere (flammable gases, fumes, vapor,
dust.) n Do not use this or any piece of test equipment without proper
training n CRT SERVICE SAFETY REMINDER : A potential danger
exists when measuring voltages in the horizontal output and damper
stages of CRT equipment. (High voltage transients greater than 6,000 V).
Refer to your CRT service manual for proper servicing instructions.
D • Warnungen und Vorsichtsmaßnahmen
n Dieses Gerät ist EN61010-1 zertifiziert für Installationsklasse III.
Anwendung ist empfohlen auf Verteilerebene und festen Anlagen sowie
untergeordneten Systemen, jedoch nicht für Starkstromnetze und
Hochspannungsanlagen. n Überschreiten Sie nie die kontinuierlichen
Überlastgrenzen der verschiedenen Meßfunktionen (siehe Spezifikationen)
– 2 –
2015.qxd 7/21/00 5:40 PM Page 4
oder andere Grenzen welche auf dem Gerät markiert sind. n Spannungen
welche am COM Eingang anliegen können sich auch auf andere Eingänge
übertragen nVorsicht beim Messen von Spannungen >20V // Strömen
>10mA // Netzstrom/-spannung bei induktiever Last oder bei Gewittern //
Strom, wenn die Sicherung durchbrennt in einem Schaltkreis mit
Leerlaufspannung >600V (>250V beim mA Eingang) // beim Messen an
Bildröhrgeräten (hohe Spannungsspitzen) n Unsersuchen Sie Gerät,
Meßkabel, Verbinder, usw. vor jeder Messung. Beschädigte Teile nicht
verwenden n Meßspitzen und Stromkreis während der Messung nicht
berühren • Sich selbst isolieren ! n Bei Strommessung, Multimeter
immer in Serie mit Schaltkreis verbinden – Nie in parallel mit
Spannungsquelle. n Sicherung immer mit gleichwertiger ersetzen. n
Gerät nicht in explosiver Umgebung verwenden.
E • Advertencias y Precauciones
n Este instrumento está homologado según EN61010-1 para la
Categoría de Instalación III. Su uso está recomendado en el nivel de
distribución y en instalaciones fijas, así como en instalaciones menores,
pero no en líneas principales de suministro, líneas aéreas ni sistemas de
cable. n No supere nunca los límites de entrada para las diferentes
funciones (vea Especificaciones), ni los límites marcados en el instrumento.
n Tensiones presentes en COM puedan estar presentes en otras entradas.
n Tenga especial cuidado al: medir tensión >20 V // corriente >10 mA //
tensión de red de CA con cargas inductivas // tensión de red de CA durante
tormentas eléctricas // corriente, si salta el fusible en un circuito con
tensión de circuito abierto >600 V (250 V en la entrada de mA) // trabajar
con pantallas TRC n Inspeccione siempre el multímetro, las puntas de
prueba, los conectores y los accesorios antes de cada uso. No utilice
ningún componente que esté dañado. n No se ponga Ud. a tierra cuando
esté tomando medidas, y no toque nunca partes expuestas de los circuitos.
n Al medir corriente, conecte siempre el multímetro EN SERIE con la carga
- NUNCA EN PARALELO con una fuente de tensión. n Nunca sustituya un
fusible con otro que no tenga las mismas especificaciones. n No utilice el
instrumento en ambientes potencialmente explosivos.
F • Avertissements et Précautions
n Cet instrument est certifié EN61010-1 catégorie d’installation III. Son
utilisation est recommandée pour le niveau de distribution de réseau, les
installations fixes et systèmes subordonnés, et non pour les installations de
puissance et lignes de transmission et câblages à haute tension. n
N’excédez jamais les limites de surcharge continues (voir spécifications) ou
d’autres limites marquées sur l’appareil. n Des tensions appliquées à
l’entrée COM peuvent être également présentes aux autres bornes d’entrée.
n Soyez très prudent quand vous mesurez des tensions >20V ou des
– 3 –
2015.qxd 7/21/00 5:40 PM Page 5
courants >10mA // tension ou courant de secteur avec charge inductive ou
par temps de tempête // du courant quand le fusible saute dans un circuit
avec tension en circuit ouvert de >600 volts (>250V pour l’entrée mA) //
dans des appareils à tube cathodique (transitoires à haute tension) n
Inspectez appareil, câbles, connecteurs avant chaque mesure. N’utilisez pas
des pièces endommagées n Ne touchez pas les pointes de touche ou le
– 4 –
3
3
/4 digit LCD with Unit, Status and Function Indicators
3
3
/
4
Digit LCD ; Dezimalpunkt, Funktions- und Einheitsanzeigen
LCD de
3
3
/
4
dígitos con indicadores de unidades, estado y función
LCD 3
3
/
4
digits ; point décimal, annonciateurs de fonctions et d’unités
40-Segment Bargraph
40-Segment Bargraf
Barra analógica
Bargraphe 40-segments
Low Battery
Batterie entladen
Pila baja
Pile déchargée
Function/Range/Off
Selector
Funktion-/Bereich-
/Aus Schalter
Selector de
Función/Escala/Off
Sélecteur fonctions/
calibres/marche-arrêt
Special Function
Buttons
Spezielle Funktions-
tasten
Teclas de función
especial
Boutons de
fonctions spéciales
High input for
voltage, resistance,
frequency,
capacitance
V HzC Eingang. Hoch
für Spannung
Widerstand,
Frequenz, Kapazität
Entrada “alta” para
tensión, resistencia y
capacidad
Entrée V HzC. Haut
pour tension,
résistance,
fréquence, capacité
COM Input – common or low input for all
measurements
COM Eingang – Referenzpunkt für alle Messungen
COM- entrada común o “baja” para todas las
medidas
Entrée COM – commun ou bas pour toutes mesures
20 A Input
20 A Eingang
Entrada 20A
Entrée 20 A
400mA Input
400mA Eingang
Entrada 400mA
Entrée 400mA
2015.qxd 7/21/00 5:40 PM Page 6
circuit pendant les mesures. Isolez-vous ! n Pour la mesure de courant,
connectez l’appareil en série avec le circuit – JAMAIS en parallèle avec une
source de tension. n Remplacez les fusibles toujours par des fusibles
équivalents n N’utilisez pas cet appareil dans des atmosphères
explosives.
PREPARATION FOR USE – UNPACKING
Your shipping carton should include the multimeter, a holster, one test
lead set (one black, one red), one 9V battery (installed), a warranty card
and this manual. If any of the items are damaged or missing, immediately
return the complete package to the place of purchase for an exchange.
D • Gebrauchsvorbereitung - Auspacken
Die Verpackung sollte enthalten: ein Multimeter, ein Meßkabelsatz (ein
schwarz, ein rot), eine 9V Batterie (im Gerät), eine Garantiekarte und diese
Anleitung. Wenn ein Teil fehlt oder beschädigt ist, bitte bei der
Verkaufsstelle umtauschen.
E • Preparación del multímetro para su uso - Desembalaje
El embalaje debe contener: el multímetro, una funda protectora, un juego de
puntas (una negra y otra roja), una pila de 9 V (instalada), una tarjeta de
garantía y este manual. Si falta algún componente u observa daños,
devuelva el conjunto al lugar donde lo adquirió para que se lo cambien.
E • Préparation pour l’Emploi - Déballage
Votre emballage doit contenir: un multimètre, un jeu de câbles de mesure
(un rouge, un noir), une pile 9V (installée), une carte de garantie et ce
manuel. Si une pièce manque ou est endommagée, retournez à votre point
de vente pur un échange.
PROTECTIVE HOLSTER
The holster/tilt stand protects the meter from accidental falls and provides
greater ease of use. The tilt stand has two positions, and a velcro-strap
allows to hang the meter from a fixture. Both test lead probes can be
attached to the holster for storage. One probe can be attached for
measurement, holding the meter with probe in one hand and the second
probe in the other hand.
– 5 –
2015.qxd 7/21/00 5:40 PM Page 7
D • Schutz-Holster
Das Schutzholster
beschirmt das Gerät
vor Stürzen und
Stößen. Der
Kippständer hat zwei
Positionen, und mit
der Schlaufe können
Sie das Gerät
aufhängen. Die Meß-
sonden können am
Holster befestigt
werden.
E • Funda Protectora
La funda con pie integrado protege el medidor en caso de caídas
accidentales. El pie inclinado tiene dos posiciones, y una banda con Velcro
permite colgar el multímetro. Las puntas de prueba pueden fijarse a la
funda.
F • Holster de Protection
Le holster protège l’appareil contre des chutes accidentelles et des chocs. Il
est muni d’une béquille à deux positions et d’une boucle de suspension.
Les sondes des cordons peuvent être attachées au holster.
OVERLOAD INDICATION
Range Overload is indicated by a flashing “4”
(MSD) and three non-flashing zeros (plus a
beep tone for voltage, capacitance and
current). Take immediate steps to remove the
cause of the overload. Select the next higher
range until a value is displayed. If overload still exists in the highest
range, remove test leads from the measurement setup as it is beyond the
range of the meter. Overload indication is normal in the OHMS and
continuity ranges when the leads are not connected to anything or when
the measured value is higher than the selected resistance range.
– 6 –
2015.qxd 7/21/00 5:40 PM Page 8
D • Überlastanzeige
Bei Bereichsüberschreitung blinkt die höchste Ziffer “4”; andere Digits
werden als “0” angezeigt. In Spannungs-, Kapazitäts- und Strombereichen
ertönt gleichzeitig ein Summton. Ursache der Bereichsüberschreitung sofort
entfernen (Höheren Bereich wählen oder Messung unterbrechen).
Überlastanzeige ist normal in Widerstands- und Durchgangsbereichen bei
offener Schaltung oder zu hohem Widerstand.
E • Indicación de Sobrecarga
La sobrecarga de escala se indica mediante el parpadeo del dígito más
significativo (“4”) y tres ceros fijos (más un bip o tono audible en tensión,
capacidad y corriente). Para eliminar la causa de la sobrecarga, seleccione
una escala más ata o interrumpa la medida. La indicación de sobrecarga es
normal en medida de OHMS y prueba de continuidad, cuando el circuito
está abierto o la resistencia es demasiado alta.
F • Indication de Surcharge
Un dépasssement de calibre est indiqué par le clignottement du chiffre le
plus significatif “4”; les autres chiffres étant “0”. Cet affichage est
accompagné d’un signal sonore dans les gammes de tension, de capacité et
de courant. Ecartez immédiatement la cause du dépassement – choisissez
une gamme plus élevée ou interrompez la mesure. L’indication de
dépassement est normale dans les gammes de résistance ou de continuité
avec circuit ouvert ou résistance trop élevée.
– 7 –
2015.qxd 7/21/00 5:40 PM Page 9
MEASURING PROCEDURES
GENERAL: Turn instrument on by turning function/range switch away
from OFF and selecting the parameter you want to measure. Six switch
positions have dual functions (resistance/continuity, diode/capacitance,
four AC/DC current ranges). How to select either is explained under the
respective function.
Ranging: This instru-
ment is autoranging on
all ranges except
current and 200mV AC.
It automatically selects
the range that gives the
best resolution for the
measured value.
“AUTO” in the display
confirms that you are in
Autoranging mode. You
can tell which range
(and function) you are
in by the position of the
decimal point and the
measurement unit displayed. In the example shown you would be in the
40V AC range. Hold a range by pressing the RANGE button. Cycle
through ranges by continuing to depress RANGE. Press and hold RANGE
button for two seconds to return to Autoranging.
Bargraph: The 40-segment bargraph indicates at which percentage of
the range you presently measure. It updates 20 times per second, and,
like an analog needle display, it is ideal for circuit adjustments.
Display Hold: When you need to concentrate on your circuit, you can
push the HOLD button to keep the reading on the display. “HLD” appears
in the display. Push the HOLD button again to resume normal display of
current value. (Automatic shut-down is disabled while “HLD” is
displayed).
General Measurement Procedures: When connecting or
disconnecting test leads to a circuit, always first turn off power to device
or circuit being tested and discharge all capacitors. Strictly observe
the max input limits. Do not change functions while test leads are
connected to circuit.
– 8 –
2015.qxd 7/21/00 5:40 PM Page 10
D • Meßprozeduren
Allgemein: Gerät einschalten durch Funktionsschalter von OFF weg, nach
einer gewünschten Meßfunktion zu drehen.
Bereichswahl: Automatisch für alle Funktionen/Bereiche, ausgenommen
Strom und 400mV AC. “AUTO” wird angezeigt. Aktuelle Funktion und
Bereich werden durch Dezimalpunkt und Meßeinheit angezeigt. Sie können
einen Bereich festhalten durch die RANGE Taste zu drücken. Sie können die
verschiedenen Bereiche manuell durchlaufen durch wiederholt RANGE zu
drücken. RANGE zwei Sekunden lang drücken um erneut automatische
Bereichswahl zu wählen.
Bargraf: Der 40-Segment Bargraf zeigt an an welcher Stelle des Bereiches
(%) Sie messen. Erneuerung: 20 pro Sekunde.
Anzeigenstop: HOLD Taste drücken um einen Meßwert auf der Anzeige
festzuhalten. “HLD” wird angezeigt. HOLD erneut drücken um aktuelle
Meßwerte anzuzeigen. Bei aktiver HOLD Funktion schaltet das Gerät nicht
automatisch ab.
Allgemeine Meßprozeduren: Beim Verbinden/Trennen der Meßkabel
mit dem/vom Schaltkreis, zuerst Schaltkreis abschalten und Kondensatoren
entladen. Eingangsgrenzen beachten. Wahlschalter nicht drehen
wenn Meßkabel mit dem Meßkreis verbunden sind.
E • Procedimientos de medida
EN GENERAL: Encienda el instrumento girando el selector de
función/escala en otra posición que no sea OFF, seleccionando el
parámetro que desee medir.
Selección de escala: Este instrumento selecciona automáticamente la
escala (auto-rango) en todas las medidas excepto corriente y 400mV AC.
Aparece la indicación “AUTO”. La escala y la función seleccionadas en cada
momento se indican mediante el punto decimal y la indicación de unidades.
Para mantener fija una escala, pulse la tecla RANGE. Para recorrer todas las
posibles escalas, pulse repetidamente RANGE. Para volver a AUTO, pulse y
mantenga pulsada RANGE durante dos segundos.
Barra analógica: La barra de 40 segmentos indica el porcentaje del
fondo de escala en el que está midiendo en cada momento. La información
se actualiza 20 veces por segundo.
Congelación de la lectura: Pulse la tecla HOLD para mantener una
lectura congelada en el visualizador. Aparece la indicación “HLD”. Pulse de
nuevo HOLD para volver a la indicación de la lectura real.
– 9 –
2015.qxd 7/21/00 5:40 PM Page 11
Procedimientos generales de medida: Antes de conectar o des-
conectar las puntas de prueba a/de un circuito , apague siempre el dispo-
sitivo o circuito sometido a prueba y descargue todos los condensadores.
Observe estrictamente los límites máximos de entrada. No cambie de
función mientras tenga las puntas de prueba conectadas a un circuito.
F • Procédures de Mesure
Général: Mettez l’appareil sous tension en tournant le sélecteur de la
position OFF vers la fonction souhaitée.
Sélection de gammes: Automatique pour toutes les fonctions et
gammes, sauf le courant et 400mV ca. “AUTO” est affiché. La fonction et la
gamme en cours sont affichées par le point décimal et l’unité. Vous pouvez
bloquer une gamme en presssant le bouton RANGE. Continuez à presser
RANGE pour parcourir les diverses gammes. Pressez RANGE pendant deux
secondes pour retourner à la sélectiion automatique.
Bargraphe: Le bargraphe à 40-segments indique à quelle position de la
gamme (%) vous mesurez. Il est raffraichi 20 fois par seconde.
Maintien de lecture: Pressez la touche HOLD pour maintenir un
affichage. “HLD” est affiché. Pressez HOLD à nouveau pour afficher les
valeurs actuelles. La coupure automatique est désactivée pendant que
HOLD est actif.
Procédures Générales: Avant de connecter/déconnecter les câbles
de mesure à/d’un ciruit, coupez l’alimentation du circuit et déchargez les
condensateurs. Ne dépassez pas les limites d’entrée. Ne tournez pas
le sélecteur tandis que les cordons sont raccordés à un circuit.
DC AND AC VOLTAGE MEASUREMENT (See Fig. 1)
Connect test leads as shown in figure 1. Turn function switch to
Vac ( ) or Vdc ( ). Touch Probe tips across voltage source (in
parallel with circuit). Voltage value will appear on Digital Display
along with the voltage polarity (DC – positive implied).
Note: The 400mV AC range needs to be selected manually. See page 8
“RANGING” for procedure.
Note for AC Measurements: Model 2015 is a True-RMS meter.
It measures the true RMS value of distorted AC voltage or current signals.
The Crest Factor handling capability is shown in table 1. The Crest Factor
is the Peak Voltage divided by the RMS voltage.
– 10 –
2015.qxd 7/21/00 5:40 PM Page 12
– 11 –
Wave form, Crest Factor
Wellenform, Crest Factor
Forme d’onde, Facteur CrŒte
Additional correction from 1.5 to 5.0
Zus tzliche Korrektur von 1.5 bis 5.0
Correction supplØm. de 1.5 5.0
Input RMS, Percentage of Full-Scale
Effektiv-Wert Eingang, % vom Endbereich
EntrØe effective, % pleine Øchelle
Table 1
Fig. 1
2015.qxd 7/21/00 5:40 PM Page 13
D • Gleich- und Wechselspannungsmessung (siehe Fig.1)
Meßkabel wie in Figur 1 gezeigt verbinden. Funktionsschalter auf
VDC ( ) oder VAC ( ) stellen. Meßspitzen mit Meßkreis
verbinden – parallel zur Spannungsquelle. Meßwert ablesen
(automatische Polaritätsanzeige bei DC Messungen).
Anmerkung: Der 400mV AC Bereich muß manuell gewählt werden. Siehe
Seite 9 “Bereichswahl” für Prozedur.
Anmerkung für AC Messungen: Modell 2015 ist ein Echt-
Effektivwertmeßgerät: Es mißt den Effektivwert von verzerrten
Spannungs- und Stromsignalen. Crest Faktor Grenzen sind in Tabelle 1
angezeigt. Der Crest Factor ist die Spitzenspannung dividiert durch die
Effektivwertspannung.
E • Medidas de tensión CC y CA (vea Fig. 1)
Conecte las puntas de prueba como se muestra en la Figura 1. Ponga
el selector de función en ACV o DCV. Toque con las puntas de prueba
los puntos de tensión (en paralelo con el circuito). Aparece el valor de la
tensión en el visualizador digital, junto con la polaridad en l caso de DCV.
Nota: La escala de 400mV AC debe seleccionarse manualmente. Consulte
el procedimiento en página 9, “Selección de escalas”.
Nota para medidas de CA: El Modelo 2015 es un medidor de
verdadero valor eficaz (TRMS). Mide el verdadero valor eficaz de
señales distorsionadas de tensión o corriente CA. En la tabla 1 se muestra
su capacidad de proceso del factor de cresta. El factor de cresta es la
tensión máxima (pico) dividida por la tensión eficaz (RMS).
F • Mesure de Tensions CC et CA (voir fig. 1)
Connectez les câbles comme indiqué en figure 1. Placez le sélecteur
sur VDC ( ) ou VAC ( ). Connectez les pointes de touche au
circuit – en parallèle avec la source de tension. Lisez la valeur.
Note: Le calibre 400mV ca doit être sélectionné manuellement. Voir page
10 “Changement de calibres” pour procédure.
Notes pour mesure en courant alternatif: Le modèle 2015 est un
appareil à mesure efficace vraie: Il mesure la valeur efficace de toutes
les formes d’ondes, avec ou sans distortion. Les limites de facteur de crête
sont indiquées dans le tableau 1. Le facteur de crête est la valeur crête
divisée par la valeur efficace.
– 12 –
2015.qxd 7/21/00 5:40 PM Page 14
DC AND AC CURRENT MEASUREMENT (See Fig. 2)
Note: the current function is not autoranging. “AUTO” does not appear in
the display, and “MANU” is displayed instead. The meter has four current
ranges, which each have their own selector switch position.
Connect red test lead to the 400mA Input for current measurements up
to 400mA or to the 20A input for current measurements to twenty
amperes (20A for 30 seconds). Connect black test lead to COM Input
Connector. Set the Function Switch to the corresponding Current
range – to A when the test lead is connected to the 20A input (mA ranges
do not work with 20A input); to 4, 40 or 400mA when test lead is
connected to the 400mA input (depending on how small a current you are
going to measure). By default, when you select a current function,
your multimeter is in AC mode. “AC” is shown in the display. Push the
yellow SHIFT button to select DC mode. “DC” is shown in the display (the
SHIFT button toggles between the two modes). Open circuit in which
current is to be measured. Securely connect test leads in series with the
load. Turn on power to circuit being tested. Read current value on
Digital Display. With 400mA input used, switch from 400mA to 40 or
4mA range if setting is too high (If you are in DC mode, the mutimeter
falls back into AC mode each time you turn the selector switch – so you
also each time need to press the yellow SHIFT button to return into DC
mode).
Note 1: Each current input has a protective fuse installed. If fuse does
blow, see page 38 for replacement information.
Note 2 – Incorrect Input Warning: The buzzer sounds when a test
lead is connected to current input, but the selector switch is not set to a
current range.
Note for AC Measurements
Model 2015 is a True-RMS meter. See Voltage Measurement section
for further details.
D • Gleich- und Wechselstrommessung (siehe Fig. 2)
Anmerkung: Die Bereichsschaltung für Strommessungen ist manuell.
“MANU” wird statt “AUTO” angezeigt.
Rotes Meßkabel für Messungen bis 400mA mit dem 400mA Eingang
verbinden, oder für Messungen bis 20A mit dem 20A Eingang (20A nur für
30 Sekunden). Schwarzes Meßkabel mit COM verbinden.
Funktionsschalter auf gewünschten Bereich stellen. Auf A wenn Meßkabel
mit 20A Eingang verbunden ist (mA Bereiche funktionieren nicht mit 20A
Eingang); auf 4, 40 oder 400mA wenn Meßkabel mit 400mA verbunden ist.
– 13 –
2015.qxd 7/21/00 5:40 PM Page 15
Strombereiche werden automatisch mit AC Funktion angewählt. “AC”
wird angezeigt. Falls gewünscht, gelbe “Shift” Taste drücken um auf DC
(Gleichstrom) überzuschalten. Meßkreis abschalten, öffnen und
Meßspitzen sicher in Serie mit dem Stromkreis verbinden. Meßkreis
einschalten. Stromwert ablesen. Mit 400mA Eingang von 400mA auf
40mA oder 4mA überschalten falls Auflösung zu niedrig ist. Wenn Sie
Gleichstrom messen, fällt das Gerät bei Bereichswechsel zurück auf
Wechselstrom. Gelbe “Shift” Taste drücken um wieder die Gleich-
stromfunktion zu wählen (“DC” Anzeige).
– 14 –
Fig. 2
2015.qxd 7/21/00 5:40 PM Page 16
Anmerkung 1: Beide Stromeingänge sind abgesichert. Sicherungen nur
mit gleichwertigen ersetzen.
Anmerkung 2 – Eingangswarnung: Wenn ein Meßkabel mit einem
Stromeingang verbunden ist, der Funktionsschalter jedoch nicht auf einer
Stromfunktion steht, wird ein akustisches Signal abgegeben. Eine solche
Messung würde das Gerät zerstören. Anmerkung für Wechsel-
strommessung: Siehe Wechselspannungsmessung.
E • Medidas de Corriente CC y CA (vea Fig. 2)
Nota: la función de corriente no tiene selección automática de escala No
aparece “AUTO” en el visualizador, sino “MANU”.
Conecte la punta de prueba roja a la entrada de 400 mA para medidas de
corriente hasta 400 mA, o a la entrada de 20A para medidas hasta 20 A.
Conecte la punta de prueba negra a la entrada COM. Ponga el selector
de función en la posición de corriente adecuada - en A cuando utilice la
entrada de 20 A (las escalas de mA no funcionan con la entrada de 20 A), y
en 4, 40 o 400 mA cuando utilice la entrada de 400 mA. Por defecto, al
seleccionar una función de corriente, el multímetro se pone en modo AC
(aparece “AC” en el visualizador). Pulse la tecla amarilla SHIFT para
seleccionar el modo DC (aparece “DC” en el visualizador). Abra el
circuito sometido a prueba y conecte las puntas de prueba en serie con la
carga. Conecte la alimentación del circuito. Lea el valor de la
corriente en el visualizador. Si utiliza la entrada de 400 mA, cambie de 400
a 40 o 4 mA si es necesario. (Si está en modo DC, el medidor pasa a modo
AC cada vez que gire el selector de función, por lo que deberá pulsar cada
vez la tecla amarilla SHIFT para volver a DC).
Nota 1: Cada entrada de corriente tiene un fusible de protección. En caso
de que salte el fusible, cámbielo por otro igual.
Nota 2 - Aviso de entrada incorrecta: El zumbador avisa si se ha
conectado la punta de prueba a una entrada de corriente y el selector no
está en una escala de corriente.
Nota para medidas de CA
El Modelo 2015 es un medidor de verdadero valor eficaz
(TRMS): Vea “Medidas de tensión”.
F • Mesure de Courant CC et CA (voir fig. 2)
Note: Les gammes de courant sont commutées manuellement.
Connectez le cordon rouge à l’entrée 400mA pour mesures jusqu’à
400mA et à l’entrée 20A pour mesures jusqu’à 20A (20A pour 30
secondes). Connectez le cordon noir à l’entrée COM. Placez le sélecteur
– 15 –
2015.qxd 7/21/00 5:40 PM Page 17
sur le calibre souhaité. Sur “A”, quand l’entrée 20A est utilisée (les gammes
mA ne fonctionnent pas avec l’entrée 20A), sur 4, 40, ou 400mA quand
l’entrée 400mA est utilisée. Une gamme de courant est automatiquement
choisie en courant alternatif. Pressez la touche jaune SHIFT pour choisir
courant continu. Ouvrez le circuit à mesurer et connectez les pointes
de touche solidement en série avec le circuit. Mettez le circuit sous
tension. Lisez la mesure. Avec l’entrée 400mA, descendez de gamme si
la résolution est insuffisante. Si vous changez de gamme en mesure de
courant continu, l’appareil se remet en courant alternatif. Pressez la touche
jaune SHIFT pour vous remettre en courant continu (DC est affiché).
Note 1: Les entrées de courant sont protégées par fusible. Ne remplacez
les fusibles que par des fusibles équivalents.
Note 2 – Avertissement d’entrée: Un signal sonore vous avertit quand
un câble de mesure est connecté à une entrée de courant, alors que le
sélecteur ne se trouve pas sur un calibre de courant. Une telle mesure
détrouirerait l’appareil.
Note pour la mesure de courant alternatif: Voir mesure de tension
alternative.
RESISTANCE MEASUREMENT (See Fig. 3)
Turn off any power to the resistance to be measured and discharge
any capacitors. Any voltage present during a resistance measurement will
cause inaccurate readings. Connect test leads as shown in figure 3.
Set Function/Range Switch to W position. Depress yellow button.
Connect test leads to circuit being measured. Read resistance value
on Digital Display. Open circuits will be displayed as an overload
condition.
D • Widerstandsmessung (siehe Fig. 3)
Stellen Sie sicher daß keine Spannung am Widerstand oder Meßkreis
anliegt. Eine Spannung würde die Messung verfälschen. Meßkabel wie
in Figur 3 verbinden. Funktionsschalter auf W Position stellen.
Gelbe Taste drücken. Meßspitzen mit Schaltkreis verbinden. Meßwert
ablesen. Offene Schaltungen oder zu hohe Widerstände werden als Überlast
angezeigt.
E • Medidas de resistencia (vea Fig. 3)
Asegúrese de que no hay tensión aplicada a la resistencia o al circuito.
Conecte las puntas de prueba como se muestra en la Figura 3. Ponga
el selector de función/escala en la posición W Conecte las puntas de
prueba al circuito sometido a medición. Lea el valor de la resistencia en
el visualizador. Un circuito abierto se indicará como sobrecarga.
– 16 –
2015.qxd 7/21/00 5:40 PM Page 18
F • Mesure de Résistance (voir fig. 3)
Assurez-vous qu’aucune tension n’est appliquée à la résistance. Une
tension fausserait le résultat. Connectez les cordons comme indiqué
en figure 3. Placez le sélecteur sur W
.
Pressez la touche jaune.
Connectez les cordons au circuit à mesurer. Lisez la valeur affichée. Un
circuit ouvert ou une résistance trop élevée est affiché comme dépassement
de calibre.
CONTINUITY TEST (See Fig. 3)
The Continuity test checks electrical continuity between two contact
points. Set the Function switch to W position. Connect black
test lead to COM input and red lead to V HzC input. Connect probe tips
to two circuit points.
The internal beeper emits a tone when resistance
is less than approx. 40 W.
– 17 –
Press yellow button to
select Resistance
Gelben Knopf dr cken um
Widerstand zu w hlen
Pulse la tecla amarilla para
seleccionar resistencia
Pressez bouton jaune pour
choisir RØsistance
Fig. 3
2015.qxd 7/21/00 5:40 PM Page 19
D • Durchgangstest (siehe Fig. 3)
Funktionsschalter auf W -Bereich stellen. Rotes Meßkabel mit
V HzC Eingang und schwarzes mit COM Eingang verbinden. Meßspizen
mit Schaltkreis verbinden. Akustisches Signal bei 100
W
.
E • Prueba de continuidad (vea Fig. 3)
Ponga el selector de función/escala en la posición W . Conecte la
punta de prueba negra a la entrada COM y la punta de prueba roja a la
entrada V HzC. Toque dos puntos del circuito con las puntas de prueba.
El zumbador suena cuando la resistencia es 100
W
.
F • Test de Continuité (voir Fig. 3)
Placez le sélecteur sur W
.
Connectez le cordon rouge à l’entrée
V HzC et le noir à l’entrée COM. Connectez les pointes de touche au
circuit. Une résistance de 40 k
W
est indiquée par un signal sonore.
DIODE TEST (See Fig. 4)
The diode test measures the forward bias of a diode junction. Connect
the test leads as shown in figure 4. Set the Function/range switch to
. Depress yellow button. Apply probe tip of red lead to the
anode and of black lead to the cathode of the diode. The meter’s display
indicates the forward voltage drop (approx. 0.6V for silicon diode or 0.4V
for germanium diode). An open diode is indicated by a large voltage
reading (>3.000 V). Reverse test lead connections to the diode to
perform a reverse bias test. A large value (>3.000 V) indicates a good
diode. Notes: A large value (>3.000 V) for both reverse and forward bias
tests indicates an open diode. A low voltage reading for both bias tests
indicates a shorted diode. If the diode is shunted by a resistor of 1000
ohms or less, it must be removed from the circuit before taking the
measurement. Bipolar transistor junctions may be tested in the same
manner described above as emitter-base and base-collector junctions are
diode junctions.
D • Dioden- und Transistortest (siehe Fig. 4)
Der Diodentest zeigt den Spannungsabfall über den Diodendurchgang
Meßkabel wie in Figur 4 verbinden. Funktionsschalter auf stellen.
Gelbe “Shift” Taste drücken. Meßkabel mit Diode verbinden – rotes mit
Anode; schwarzes mit Kathode. Spannungsabfall in Durchlaßrichtung
ablesen (ung. 0.600V für eine Silikon-Diode und 0.400V für eine Ger-
– 18 –
2015.qxd 7/21/00 5:40 PM Page 20
maniumdiode. Eine offene Diode wird mit einem hohen Wert angezeigt
(>3.000V). Verbindung umdrehen um in Sperrrichtung zu messen. Eine
hohe Ablesung (>3.000V) zeigt eine gute Diode an.
Anmerkung: Eine hohe Ablesung (>3.000V) in beiden Richtungen zeigt
eine offene Diode an; eine niedrige Ablesung eine kurzgeschlossene Diode.
Transistorübergänge können wie Dioden getestet werden.
E • Comprobación de diodos y transistores (vea Fig 4)
En esta prueba se mide la polarización directa de la unión de un diodo.
Conecte las puntas de prueba como se muestra en la Figura 4. Ponga el
selector de función/escala en la posición Pulse la tecla amarilla.
Aplique la punta de prueba roja al ánodo y la negra al cátodo. El
visualizador indica la caída de tensión directa (aprox. 0.6 V para diodos de
Si, o 0.4 V para diodos de Ge). Una unión abierta se indica mediante una
lectura muy elevada de tensión (>3.000 V). Invierta la conexión de las
puntas de prueba para comprobar la polarización inversa del diodo. Un
valor elevado (>3.000 V) indica un diodo en buenas condiciones.
– 19 –
Fig. 4
2015.qxd 7/21/00 5:41 PM Page 21
Notas: Un valor elevado (>3.000 V) en ambos sentidos de polarización
indica un diodo abierto. Un valor bajo en ambos sentidos indica un diodo
cortocircuitado. Las uniones de un transistor bipolar son diodos, por lo que
pueden comprobarse como tales.
F • Test de Diodes et de Transistors (voir fig. 4)
Le test de diodes affiche la chute de tension à travers le transistor.
Connectez les cordons comme indiqué en figure 4. Placez le sélecteur
sur . Pressez la touche jaune SHIFT. Connectez les pointes de
touche à la diode – le rouge à l’anode, le noir à la cathode. Lisez la chute
de tension en direction passante (environ 0.600V pour une diode au Si;
0.400V pour une diode au Ge. Une diode ouverte est affichée par une
lecture élevée (>3.000V). Inversez la connection pour mesurer en
direction de bloquage. Une bonne diode est affichée par une lecture élevée
(>3.000V).
Notes: Une lecture élevée dans les deux directions indique une diode
ouverte; une lecture basse indique une diode court-circuitée. Les jonctions
de transistors peuvent être testées comme des diodes.
CAPACITANCE MEASUREMENTS (See Fig. 5)
Discharge all voltage from the capacitor before measurement. A safe way
to discharge a capacitor is to connect a 100 kW resistor across the two
capacitor leads. Connect test leads as shown in figure 5. Set the
Function/Range switch to the range. When measuring
electrolytic capacitors, observe correct polarity (connect the COM test
lead to the negative capacitor lead marked –). Read the capacitance
value on the display.
D • Kapazitätsmessung (siehe Fig 5)
Kondensator entladen (über 100 k
W
Widerstand). Meßkabel wie in Figur
5 verbinden. Funktionsschalter auf stellen. Bei
elektrolytischen Kondensatoren Polarität beachten. Meßwert ablesen.
E • Medidas de Capacidad (vea Fig. 5)
Descargue la tensión presente en el condensador (a través de una
resistencia de 100 K
W
). Conecte las puntas de prueba como se muestra
en la Figura 5. Ponga el selector de función/escala en la posición
. (Tenga en cuenta la polaridad di va a medir condensadores
electrolíticos). Lea el valor de la capacidad en el visualizador.
– 20 –
2015.qxd 7/21/00 5:41 PM Page 22
F • Mesure de Capacité (voir fig. 5)
Déchargez le condensateur (à travers une résistance de 100k
W
).
Connectez les cordons comme indiqué en figure 5. Placez le sélecteur
sur (respectez la polarité pour les condensateurs électrolytiques).
Lisez la valeur.
FREQUENCY MEASUREMENT (See Fig. 6)
Connect the red test lead to the VW HzC Input and the black test lead to
the COM Input. Set the Function/Range switch to Hz position.
Connect test lead tips across the signal source
. Read the frequency
value on the display. Note: Frequency counter operates in a 1000 count
mode. Ignore numerals above decimals.
D • Frequenzmessung (siehe Fig. 6)
Rotes Meßkabel mit V
W
HzC Eingang verbinden und schwarzes mit
COM Eingang. Funktionsschalter auf Hz stellen. Meßspitzen mit
Signalquelle verbinden. Meßwert ablesen. Anmerkung: Frequenz-
messung erfolgt in einem 1000-Punkte Anzeigenbereich. Die Ziffern über
den Dezimalpunkten sind nicht zu beachten.
– 21 –
Fig. 5
2015.qxd 7/21/00 5:41 PM Page 23
E • Medidas de Frecuencia (vea Fig. 6)
Conecte la punta de prueba roja a la entrada V
W
HzC y la negra a la
entrada COM. Ponga el selector de función/escala en la posición Hz.
Conecte las puntas de prueba a la fuente de señal. Lea el valor de la
frecuencia en el visualizador. Nota: El medidor de frecuencia opera en un
modo de 1000 cuentas. Ignore las cifras por encima de los decimales.
F • Mesure de Fréquence (voir fig. 6)
Connectez le cordon rouge à l’entrée V
W
HzC et le noir à l’entrée COM.
Placez le sélecteur sur Hz. Connectez les pointes de touche à la source de
fréquence. Lisez la valeur. Note: Pour la fréquence, l’affichage est limité à
1000 points. Les chiffres au-dessus des points décimaux sont à ignorer.
– 22 –
Fig. 6
2015.qxd 7/21/00 5:41 PM Page 24
MIN/MAX MEASUREMENT
Press MN/MX button to enter
MIN Recording mode The
meter switches to manual
ranging with HOLD invoked.
The last min. value is kept on
the display until replaced by a
lower reading. Press MN/MX
button again to change to
MAX recording mode. (MIN
or MAX annunciator indicates
what value is recorded).
Overrange (flashing 4) is
displayed when the range is
exceeded. The bargraph continues to follow the measured signal. You can
press the HOLD-button to stop recording, and press HOLD again to
restart it. If recording is stopped, the minimum, maximum or present
value as well as the bargraph are frozen. During MIN/ MAX recording
mode Automatic Shutdown is disabled. With HOLD released, press
RANGE button for two seconds to clear MX/MN and to return to
autoranging mode.
D • Min/Max Messung
MN/MX Taste drücken um MIN Meßfunktion zu wählen. Das Multimeter
schaltet nach manueller Bereichswahl mit aktivierter Anzeigesperre (HOLD).
Die letzte niedrigste Messung wird auf der Anzeige festgehalten bis er durch
einen niedrigeren Wert ersetzt wird. Erneut MN/MX Taste drücken um
Maximale Werte festzuhalten. (MIN oder MAX Funktionsanzeige gibt an
welcher Wert gemessen wird.) Bei Bereichsüberschreitung wird Überlast
angezeigt. Während die Anzeige fesgehalten wird, folgt der Bargraf dem
gemessenen Signal. Mit der HOLD Taste können Sie die Aufwertung
stoppen und wieder starten (Min, Max, laufender Wert und Bargraf werden
festgehalten). Während der MIN oder MAX Messung schaltet das Gerät
nicht automatisch ab. Mit gelöster HOLD Taste die RANGE Taste zwei
Sekunden lang drücken um die MX/MN Funktion zu verlassen und nach
automatischer Bereichsschaltung zurückzukehren.
E • Medidas Min/Max
Pulse la tecla MN/MX para entrar en el modo de registro MIN. El medidor
entra en selección manual de escala, con HOLD activada. El visualizador
mantiene el último valor mínimo hasta que se registre un valor inferior.
– 23 –
2015.qxd 7/21/00 5:41 PM Page 25
Pulse de nuevo MN/MX para cambiar al modo MAX. (El indicador
MIN/MAX señala qué valor se está midiendo). Aparece la indicación de
sobrecarga si se supera el fondo de escala. La barra analógica continúa
siguiendo a la señal medida. Pulse de nuevo la tecla MN/MX para visualizar
las medidas reales (el medidor permanece en modo de escala manual).
Puede pulsar la tecla HOLD para detener el registro, y nuevamente HOLD
para reanudarlo. Si se detiene el registro, tanto el valor mínimo, máximo o
actual como la barra analógica quedan congelados. Durante el modo de
registro MIN/MAX se anula la función de apagado automático. Tras liberar
HOLD, pulse la tecla RANGE durante 2 segundos para cancelar MX/MN y
volver al modo de escala automática.
F • Mesure Min/Max
Pressez la touche MN/MX pour choisir la fonction MIN. Le multimètre se
met en sélection manuelle avec maintien de lecture (HOLD) activé. La
dernière mesure la plus basse est gardée sur l’afficheur jusqu’à être
remplacée par une nouvelle valeur inférieure. Pressez à nouveau la touche
MN/MX pour afficher et maintenir le valeur la plus élevée. (MIN ou MAX est
affiché selon la valeur mesurée.) Pendant que la valeur minimale ou
maximale est affichée, le bargraphe suit le signal actuel. Avec la touche
HOLD, vous pouvez arrêter la mesure et la résumer (Min, Max, la valeur
courante et le bargraphe sont gelés). Pendant l’enregistrement MIN ou
MAX, la coupure automatique est désactivée. Avec HOLD désactivé, pressez
RANGE pendant deux secondes pour quitter la fonction MX/MN et retourner
vers la sélection automatique.
RELATIVE MODE
With a measured value on the
display, press REL to enter the
Relative Mode. “REL D
appears on display, the
displayed reading is memo-
rized as reference value and
the display resets to zero. For
the next measurements, the
display shows the difference
between the new value and the
stored reference value. Range
Lock also works in Relative
Mode. Press and hold down
the REL-button for 2 seconds to exit the relative mode, or press RANGE
for two seconds to clear REL and return to autoranging.
– 24 –
2015.qxd 7/21/00 5:41 PM Page 26
D • Relativ-Messung
Mit einem Meßwert auf der Anzeige, REL Taste drücken um Relativ-
Messung zu wählen. “REL
D
” wird angezeigt. Der angezeigte Wert wird als
Referenzwert gespeichert und die Anzeige wird auf Null gesetzt. Für
folgende Messungen wird der Referenzwert vom gemessenen Wert
abgezogen und die Differenz angezeigt. REL Taste oder RANGE Taste zwei
Sekunden lang drücken um Relativ-Messung zu verlassen.
E • Medidas Relativas
Teniendo una medida en el visualizador, pulse REL para entrar en el modo
de medidas relativas. Aparece “REL
D
“ en el visualizador, memorizándose
la lectura presente como valor de referencia y reponiéndose el visualizador
a cero. Para las medidas subsiguientes, el visualizador indica la diferencia
entre el nuevo valor y el almacenado como referencia. En modo Relativo
también funciona el bloqueo de escalas. Pulse y mantenga pulsada la tecla
REL durante 2 segundos para salir del modo relativo, o bien RANGE
durante 2 segundos para cancelar REL y volver al modo de escala
automática.
F • Mesure Relative
Avec une mesure sur l’afficheur, pressez la touche REL. “REL
D
” est affiché.
La valeur affichée est prise comme référence, et l’affichage est mis à zéro.
Pour les prochaines mesures, cette valeur-référence est déduite du résultat,
et seule la différence est affichée. Pressez la touche REL pendant deux
secondes pour quitter la mesure relative.
AUTOMATIC SHUTDOWN
In order to save battery life, your multimeter turns off automatically after
30 minutes of inactivity. You can turn it back on by simply pressing the
RESET button. You can disable the automatic shutdown by pressing and
holding the RESET button while you initially switch on the instrument by
turning the selector switch away from OFF to a desired measurement
function.
D • Automatische Abschaltung
Ihr Multimeter schaltet nach 30 Minuten Inaktivität automatisch ab. Sie
können es wieder einschalten durch die gelbe RESET Taste zu drücken. Sie
können die automatische Abschaltung ausschalten durch beim Einschalten
des Multimeters (Drehen des Funktionsschalters weg von OFF nach einer
gewünschten Funtion) gleichzeitig die RESET Taste zu drücken.
– 25 –
2015.qxd 7/21/00 5:41 PM Page 27
E • Apagado Automatico
El multímetro se apaga automáticamente cuando transcurren 30 minutos
sin actividad. Para volver a encenderlo basta con pulsar la tecla RESET. Es
posible desactivar la función de apagado automático, pulsando y
manteniendo pulsada RESET al mismo tiempo que enciende el instrumento,
girando el selector a la posición deseada.
F • Coupure Automatique
Votre multimètre s’éteint automatiquement apres 30 minutes d’inactivité.
Vous pouvez le ralummer en pressant la touche jaune RESET. Vous pouvez
désactiver la coupure automatique en pressant la touche RESET au moment
de mettre votre mutlimètre sous tension (tourner le sélecteur de OFF vers
une fonction souhaitée).
INCORRECT INPUT WARNING
The Input Warning (beeper) is a feature to protect the meter and you from
unintentional misuse. It warns you when a test lead is placed in a current
input jack and the Selector switch is not set to current. (If a DMM is
connected to a voltage test point with leads set for current, very high
current could result). The warning reminds you that a test lead needs to
be changed from a current jack to the voltage jack. All current ranges are
fused with fast acting fuses.
D • Eingangswarnung
Ein akustisches Signal warnt Sie wenn ein Meßkabel mit einem
Stromeingang verbunden ist, der Funktionsschalter jedoch nicht auf Strom
steht. Eine solche Messung würde das Gerät zerstören.
E • Aviso por Conexion Incorrecta de las Puntas de Prueba
El zumbador suena si ha conectado una punta de prueba a una entrada de
corriente y el selector no está en una escala de corriente. De seguir adelante
con la medida podría dañar el multímetro.
F • Avertissement d’Entrée
Un signal sonore vous avertit quand un cordon de mesure est connecté à
une entrée de courant alors que le sélecteur n’est pas placé sur une gamme
de courant. Une telle mesure endommagerait l’appareil.
– 26 –
2015.qxd 7/21/00 5:41 PM Page 28
SAFETY TEST LEADS
The test leads included with your meter have shrouded banana plugs to
eliminate the possibility of shock if the plugs accidentally pull out of the
meter while making a measurement. The test leads also have insulated
probe tips to avoid causing shorts when making measurements in dense
component areas. Replacement part number for safety test leads is TL245.
D • Sicherheitsmeßkabel
Die Meßkabel haben versenkte Bananenstecker um elektrischen Schock zu
vermeiden. Die Meßspitzen sind zum Teil isoliert, um Kurzschlüsse in
dichten Schaltungen zu vermeiden. Diese Isolation kann entfernt werden.
E • Puntas de Prueba de Seguridad
Las puntas de prueba suministradas con el multímetro incluyen unos
conectores de banana protegidos para eliminar la posibilidad de descargas
eléctricas. Las puntas están parcialmente aisladas para evitar cortocircuitos
en áreas con alta densidad de componentes. El usuario puede quitar dicho
aislante si lo desea.
F • Cordons de Sécurité
Les fiches banane des cordons sont munis de protecteurs fixes afin de
supprimer les risques de chocs électriques. Les pointes métalliques sont
partiellement isolées pour éviter des court-circuits dans des circuits
denses. Cette isolation peut être enlevée.
– 27 –
2015.qxd 7/21/00 5:41 PM Page 29
General Specifications
Display: 3-3/4 digit LCD, 4000
counts, 9999 counts in Frequency
mode, 40 segment bargraph with
function and unit annunciators.
Polarity Indication: Automatic,
positive implied, negative (–)
indicated
Zero Adjustment: Automatic
Overrange Indication: Most significant
digit (MSD) blinks. Other digits
shown as zero.
Low Battery Indication: . Change
battery immediately.
Display Update Rate: 2/sec, nominal;
1/sec in capacitance and frequency
modes; 20/sec for bargraph.
Operating Temp.: 0 to 50°C, 0 to 70%
Relative Humidity
Storage Temp: -20°C to 60°C, 0 to
80% RH, battery removed
Temperature coefficient: 0.1 x (spec.
accuracy)/°C (<18°C or >28°C°
Power: Standard 9-volt battery, NEDA
1604, JIS 006P, IEC 6F22
Auto Power-Off: Meter turns off after
30 minutes of inactivity.
Battery Life (Typical): 500 hours
alkaline, 250 hours carbon-zinc
Dimensions (H x W x D): 7.5” x 3.5” x
1.3”, (175 x 90 x 34 mm)
Weight (incl. battery): 568 g (20 oz)
Accessories: Test leads, spare fuse,
battery, and Operator’s Manual
Safety: Meets EN61010-1 Cat III
1000V; EMC: Meets EN50081-1,
EN50082-2.
EMC: This product
complies with require-
ments of the following European
Community Directives:
89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC
(Low Voltage) as amended by
93/68/EEC (CE Marking).
However, electrical noise or
intense electromagnetic fields in
the vicinity of the equipment may
disturb the measurement circuit.
Measuring instruments will also
respond to unwanted signals that
may be present within the
measurement circuit. Users
should exercise care and take
appropriate precautions to avoid
misleading results when making
measurements in the presence of
electronic interference.
Electrical Specifications
Accuracies at 23°C ± 5°C, <75% RH.
DC Volts
Ranges: 400mV, 4, 40, 400, 1000V
Resolution, 100µV in 400mV range:
Accuracy, all Rgs: ±(0.25%rdg +2dgt)
Input Impedance, 400mV rng.
>1000 MW ; 4V rng: 11 MW ; 40V-
1000V rngs: 10 M W
OL Protection: 1100VDC or Peak AC.
6kV transient protection.
AC Volts (50-1000Hz).
Model 2015: True RMS meter. Accuracies
apply to all waveforms.
Ranges: 400mV, 4, 40, 400, 750V
Resolution: 100µV in 400mV range
Crest Factor 3:1
Accuracy ±(%rdg +nr digits):
Range 50-500Hz 500-1000Hz
400mV* ±(1.0% +5) unspecified
40V-400V ±(0.8% +3) ±(1.0% +5)
All other ±(0.8% +3) ±(1.2% +5)
Input Impedance,
400mV rg. >1000 MW ; 4V rg:
11 MW ; 40V-750V rgs: 10 M W
OL Protection: 1100VDC or Peak AC.
6kV transient protection.
– 28 –
SPECIFICATIONS
2015.qxd 7/21/00 5:41 PM Page 30
* Frequency response 50-100Hz
DC Current
Ranges: 4, 40, 400mA, 20A
Resolution, 4mA range: 1µA
Accuracy, 4mA to 400mA ranges:
±(0.8%rdg + 2dgt)
20A range: ±(1.2%rdg ±4dgt)
Voltage burden: 500mV.
400mA rg.: 900mV
OL Protection: mA input: 0.5A/250V
fast blow ceramic fuse (FP250,
6.4x25.4mm); 10kA interrupt rating
20A input: 10A continuous, 20A for
30 seconds. Fuse: Littel Fuse KLK
20A or Bussman KTK 20A
10.3x38.1mm fast acting fuse.
100kA interrupt rating.
AC Current (50-1000Hz)
Model 2015: True RMS meter. Accuracies
apply to all waveforms.
Ranges: 4, 40, 400mA, 20A
Resolution, 4mA range: 1µA
Crest Factor 3:1
Accuracy, 4mA to 400mA ranges:
±(1.2%rdg + 4dgt)
20A range: ±(1.5%rdg ±5dgt)
Voltage burden: see DC current
OL Protection: see DC current.
Resistance
Range/Accuracy:
400
W
/±(0.8%rdg + 4dgt)
4 k- 400 kW /±(0.7%rdg + 2dgt);
4 M
W /±(1.0%rdg + 4dgt);
40 M
W/±(1.5%rdg + 4dgt)
Resolution, 400
W range: 100 m W
Test voltage, all ranges: 0.4VDC
OL protection, all Ranges: 500VDC
or AC RMS
Diode Test
Range: 4V
Resolution: 1mV
Test current: 0.6mA
Display: Forward Junction Voltage
Accuracy: ±(1.0%rdg +1dgt)
Open Circuit Voltage: 3.2Vdc
Overload prot.: 500VDC or RMS AC
Continuity Indicator
Range: 400 Ohms
Audible threshold: < 40 W
Response time: approx. 100 ms
Open circuit voltage: 400 mV DC
OL Protection: 500 VDC or RMS AC
Capacitance
Ranges: 4, 40, 400nF, 4, 40µF
Accuracy, 4 nF range: ±(1.0%rdg +
40dgt); 40 nF-4µF: ±(1.0%rdg +
4dgt); 40 µF (<20µF): ±(1.0%rdg +
4dgt);
40 µF (>20 µF): ±(5.0%rdg + 8dgt)
Test frequency, all ranges: 1Hz
Note: use relative mode for 4nF and 40nF
ranges.
Resolution, 4nF range: 0.001nF
OL Protection: 500 VDC or RMS AC
Frequency Counter
Ranges: 100 Hz, 1, 10, 100, 400 kHz
Accuracy, 100 Hz rg: ±(0.1%rdg
+10dgt);
1 & 10 kHz: ±(0.1%rdg +4dgt);
100 kHz: ±(0.1%rdg +8dgt);
400 kHz: ±(0.1%rdg +20dgt)
Resolution, 100Hz range: 0.01Hz
Trigger Level - 100 Hz-100 kHz:
40 mV; 100kHz-400kHz: 400mV
OL Protection: 500VDC or RMS AC
Note: For frequencies below 1Hz, display
indicates 00.00Hz.
Optional Accessories
RF241 650MHz RF Probe
DL243C Basic Test Lead Set
DL248C Deluxe Test Lead Set
TL253A Temperature Converter
(900°C/1652°F)
TL245 Replacement Safety Test Leads
CT231A 150A AC Current Clamp
CT232A 1000A AC Current Clamp
CT234A 400A AC Current Clamp
– 29 –
2015.qxd 7/21/00 5:41 PM Page 31
CT235 1000A AC/DC Clamp
CT236A 500A AC Clamp (mV output)
CT237 200A AC/DC Current Clamp
CT238 20A AC/DC Current Clamp
VC221A Padded Vinyl Case. Fits meter with
holster
VC231 Padded Vinyl Case. Fits meter
without holster.
DC205C Deluxe Hard-Shell Carry Case
DC207C Large Deluxe Hard-Shell Carry
Case with extra space for
accessories
– 30 –
D SPEZIFIKATIONEN
Allgemeine Spezifikationen
Anzeige: 3-3/4 Digit LCD, 4000
Punkte, 9999 Punkte bei
Frequenzmessung, 40-Segment
Bargraf mit Funktions- und
Einheitsanzeigen.
Polaritätsanzeige: Automatisch,
positiv unterstellt, negativ (–)
angezeigt.
Nullabgleich: Automatisch
Überlastanzeige: Höchste Ziffer (4)
blinkt. Andere als Null angezeigt.
Entladene Batterieanzeige: .
Batterie sofort wechseln.
Anzeigenauffrischung: 2/Sek,
nominal; 1/Sek für Kapazität und
Frequenz; 20/Sek für Bargraf.
Betriebstemp.: 0 bis 50°C, 0 bis 70%
Relative Feuchtigkeit
Lagertemp: -20°C bis 60°C, 0 bis
80% R.F., Batterie entfernt.
Temperaturkoeffizient: 0.1 x
(angegebene Genauigkeit)/°C
(<18°C oder >28°C°
Stromversorgung: Standard 9-Volt
Transistorbatterie, NEDA 1604, JIS
006P, IEC 6F22
Automatische Abschaltung: Nach 30
Minuten Inaktivität.
Autonomie (typisch): 500 Stunden
Alkaline, 250 Stunden Zinn-Kohle
Abmessungen (H x B x T): 175 x 90 x
34 mm
Gewicht mit Batterie: 568 g
Zubehör: Meßkabel, Ersatzsicherung,
Batterie, Anleitung
Sicherheit: Gemäß EN61010-1 Cat.III
1000V; EMC: Gemäß EN50081-1,
EN50082-2
EMC Dieses Produkt beant-
wortet an die Bestimmungen
der folgenden EWG
Richtlinien: 89/336/EEC
(Elektromagnetische Kompatibilität)
und 73/23/EEC (Niedrige Spannung)
geändert durch 93/68/EEC (CE
Marking).
Elektrisches Rauschen und starke
magnetische Felder in der direkten
Umgebung des Meßgerätes können
jedoch den Meßkreis beeinflussen.
Das Gerät kann auch durch Störsig-
nale im gemessenen Schaltkreis
beeinflußt werden. Der Anwender muß
Vorsichtsmaßnahmen treffen um
irreführende Meßergebnisse bei
Messungen in der Umgebung von
starken elektromagnetischen Feldern
zu vermeiden.
Elektrische Spezifikationen
Genauigkeit bei 23°C ± 5°C, <75% R.F.
Gleichspannung
Bereiche: 400mV, 4, 40, 400, 1000V
Auflösung, 100µV im 400mV Ber.
Genauigkeit, alle Bereiche:
±(0.25%vMW +2Dgt)
Eingangsimpedanz, 400mV Bereich
>1000MW, 4V Ber: 11M W; 40V-
1000V Ber.: 10MW
Überlastschutz: 1100VDC oder AC
Spitze. 6kV Transienten.
Wechselspannung (50-1000Hz).
Modell 2015: Echt-Effektivwert Messung.
Bereiche: 400mV, 4, 40, 400, 750V
2015.qxd 7/21/00 5:41 PM Page 32
Auflösung: 100µV im 400mV Ber.
Crest Faktor: 3:1
Genauigk. ±(%Bereich +Anz. Digits):
Bereich 50-500Hz 500-1000Hz
400mV* ±(1.0% +5) nicht spezifiz.
40-400V ±(0.8% +3) ±(1.0% +5)
Andere ±(0.8% +3) ±(1.2% +5)
Eingangsimpedanz, 400 mV Ber.
>1000 M W; 4V Ber.: 11 MW ; 40V-
750V Ber.: 10 MW
Überlastschutz: 1100VDC oder AC
Spitze. 6kV Transienten.
* Frequenzgang: 50-100Hz
Gleichstrom
Bereiche: 4, 40, 400mA, 20A
Auflösung, 4mA Bereich: 1µA
Genauigkeit, 4mA bis 400mA
Bereiche: ±(0.8%vMW + 2Dgt)
20A Bereiche: ±(1.2%vMW ±4Dgt)
Spannungsabfall: 500mV, ausg.
400mA Bereich.: 900mV
Überlastschutz: mA Eingang:
0.5A/250V schnelle Sicherung
(FP250 - 6.4 x 25.4mm); 10kA
Trennvermögen
20A Eingang: 10A kontinuierlich,
20A für 30 Sekunden.
Littel Fuse
KLK 20A oder Bussman
KTK 20A
10.3x38.1mm flinke Sicherung.
100kA Trennvermögen.
Wechselstrom (50-1000Hz)
Modell 2015: Echt-Effektivwert Messung.
Bereiche: 4, 40, 400mA, 20A
Auflösung, 4mA Bereich: 1µA
Crest Faktor: 3:1
Genauigkeit, 4mA bis 400mA
Bereiche: ±(1.2%vMW + 4Dgt)
20A Bereich: ±(1.5%vMW ±5Dgt)
Spannungsabfall: siehe Gleichstrom
Überlastschutz: siehe Gleichstrom
Widerstand
Bereich/Genauigkeit:
400
W
/±(0.8%vMW + 4Dgt)
4k-400 kW/±(0.7%vMW + 2Dgt);
4 M
W/±(1.0%vMW + 4Dgt);
40M
W/±(1.5%vMW + 4Dgt)
Auflösung, 400
W Bereich: 100mW
Test Spannung, alle Bereiche: 0.4V
Überlastschutz, alle Bereiche:
500VDC oder AC effektiv
Diodentest
Bereich: 4V
Aulösung: 1mV
Teststrom: 0.6mA
Anzeige: Spannungsabfall in
Durchlaßrichtung
Genauigkeit: ±(1.0%vMW +1Dgt)
Leerlaufspannung: 3.2VDC
Überlastschutz: 500VDC oder AC
effektiv
Durchgangstest
Bereich: 400 Ohms
Akustisches Signal bei: < 40 W
Ansprechzeit: ung. 100ms
Leerlaufspannung: 400mV DC
Überlastschutz: 500VDC oder RMS
effektiv
Kapazität
Bereiche: 4-400nF, 4-40µF
Genauigkeit, 4nF Bereich:
±(1.0%vMW + 40Dgt); 40nF-4µF:
±(1.0%vMW + 4Dgt); 40µF
(<20µF): ±(1.0%vMW + 4Dgt);
40µF (>20µF): ±(5.0%vMW + 8Dgt)
Test Frequenz, alle Bereiche: 1Hz
Anmerkung: Relativmessung für 4nF und
40nF Bereiche verwenden
Auflösung, 4nF Bereich: 0.001nF
Überlastschutz: 500VDC oder AC
effektiv
Frequenzzähler
Bereiche: 100Hz, 1, 10, 100, 400kHz
Genauigkeit, 100Hz Bereich:
±(0.1%vMW +10Dgt);
1 & 10kHz: ±(0.1%vMW +4Dgt);
100kHz: ±(0.1%vMW +8Dgt);
400kHz: ±(0.1%vMW +20Dgt)
– 31 –
2015.qxd 7/21/00 5:41 PM Page 33
Auflösung, 100Hz Bereich: 0.01Hz
Überlastschutz: 500VDC oder AC
effektiv
Anmerkungen: Für Frequenzen unter 1Hz,
wird 00.00Hz angezeigt.
Optionszubehör
RF241 650MHz Hochfrequenzsonde
DL243C Standard Testkabelsatz
DL248C Deluxe Testkabelsatz
TL253A Temperaturumsetzer
(900°C/1652°F)
TL245 Ersatz-Sicherheitsmeßkabel
CT231A 150A Wechselstromzange
CT232A 1000A Wechselstromzange
CT234A 400A Wechselstromzange
CT235 1000A AC/DC Stromzange
CT236 500A Wechselstromzange (mV
Ausgang)
CT237 200A AC/DC Stromzange
CT238 20A AC/DC Stromzange
VC221A Gefütterte Vyniltasche. Für Meter
und Holster
VC231 Gefütterte Vyniltasche. Für Meter
ohne Holster
DC205C Deluxe-Hartmantel-Tragetasche
DC207C Große Deluxe-Hartmantel-
Tragetasche mit mehr Platz
r
Zubeh
ö
r
– 32 –
E ESPECIFICACIONES
Especificaciones Generales
Visualizador: LCD de 3-3/4 dígitos,
4000 cuentas, 9999 cuentas en
medidas de frecuencia, barra
analógica de 40 segmentos, indi-
cadores de función y unidades
Indicación de polaridad: Automática,
positiva implícita, negativa indicada
Ajuste de cero: Automático
Indicación de sobrecarga: Dígito más
significativo intermitente, otros
dígitos fijos a cero
Indicación de “pila baja”: . Esta
indicación señala el momento de
cambiar la pila.
Frecuencia de refresco de la lectura: 2
veces/segundo nominal; 1 vez/s en
medidas de capacidad y frecuencia;
20 veces/s la barra analógica.
Temp. de funcionamiento: 0 a 50 °C,
0 a 70% H.R.
Temp. de almacenamiento: -20 a 60
°C, 0 a 80% H.R., sin pila.
Coef. de temperatura: 0.1 x (precisión
espec.)/°C (<18 °C o >28 °C)
Alimentación: Pila normal de 9 V,
NEDA 1604, JIS 006P, IEC 6F22
Apagado automático: El medidor se
apaga transcurridos 30 minutos sin
actividad
Duración de la pila (típica): alcalina
500 hrs, carbono-zinc 250 hrs
Dimensiones: 175 x 90 x 34 mm
Peso (pila incluida): 568 g
Accesorios: Puntas de prueba, fusible
de repuesto, pila y Manual de
Instrucciones
Seguridad: Según normas EN61010-1
Cat.III 1000V; EMC: Según
EN50081-1, EN50082-2
EMC: Este producto cumple
los requisitos de las
siguientes Directivas de la Comunidad
Europea: 86/336/EEC (Compatibilidad
Electromagnética) y 73/23/EEC (Baja
Tensión), con enmiendas según
93/68/EEC (Marcado CE).
No obstante, la presencia de ruido
eléctrico o campos electromagnéticos
intensos en las proximidades del
equipo pueden introducir
perturbaciones en los circuitos de
medida. Los instrumentos de medida
también responden a las señales no
deseadas que puedan estar presentes
en los circuitos de medida. El usuario
deberá tomar las precauciones
necesarias para evitar obtener
resultados incorrectos cuando realiza
medidas en presencia de
interferencias electromagnéticas.
2015.qxd 7/21/00 5:41 PM Page 34
Especificaciones eléctricas
Valores precisión a 23 ºC ±5 ºC, H.R. <75%
Voltios CC
Escalas: 400 mV; 4, 40, 400, 1000 V
Resolución: 100 mV en esc. de 400 mV
Precisión (todas las escalas):
±(0.25% lect. +2 dígitos)
Imped. de entrada: Esc. 400 mV:
>1000MW ; escala 4 V: 11MW;
escalas 40-1000 V: 10MW
Protecc. sobrec.: 1100 V CC o pico
CA. Protección transitorios 6 KV
Voltios CA (50 - 1000 Hz)
Modelo 2015: Medidor de verdadero
valor eficaz - precisión aplicable a todas
las formas de onda
Escalas: 400 mV; 4, 40, 400, 750 V
Resolución: 100 mV en esc. de 400 mV
Factor Cresta: 3:1
Precisión: ±(% escala +nº de dígitos):
Escala 50-500 Hz 500-1000 Hz
400 mV* ±(1.0% +5) sin especificar
40-400 V ±(0.8% +3) ±(1.0% +5)
Otras ±(0.8% +3) ±(1.2% +5)
Impedancia de entrada: escala 400mV:
>1000 MW ; escala 4 V: 11 MW;
esc. 40-750 V: 10 MW
Protecc. sobrec.: 1100 V CC o pico
AC. Protección transitorios 6 KV
*Respuesta en frecuencia 50-100 Hz
Corriente CC
Escalas: 4, 40, 400 mA; 20 A
Resolución: 1 mA, escala de 4 mA
Precisión: escalas 4 mA- 00 mA:
±(0.8% lect +2 dgt); escala 20 A:
±(1.2% lect ±4 dgt)
Lastre de tensión: 500 mV. Escala 400
mA: 900 mV
Protección sobrecorriente: Entrada
mA: fusible rápido de 0.5A/250V
(FP250 - 6.4 x 25.4mm); nivel de
interrupción 10 KA
Entrada A: 10 A continuo, 20 A
durante 30 s. Fusible cerámico
rápido Littel Fuse KLK 20A o Buss-
man KTK 20A 10.3x 38.1mm. Nivel
de interrupción 100 KA.
Corriente CA (50 - 1000 Hz)
Modelo 2015: Medidor de verdadero
valor eficaz - precisión aplicable a todas
las formas de onda
Escalas: 4, 40, 400 mA; 20 A
Resolución: 1 mA en la esc. de 4 mA
Factor Cresta: 3:1
Precisión: escalas 4 mA a 400 mA:
±(1.2% lect. +4 dgt); escala 20 A:
±(1.5% lect ±5 dgt)
Lastre de tensión: vea Corriente CC
Protecc. sobrec.: vea Corriente CC
Resistencia
Escalas/Precisión:
400 W /±(0.8% lect +4 dgt)
4 k-400 K W/±(0.7% lect +2 dgt)
4 M W/±(1.0% lect +4 dgt)
40 M W/±(1.5% lect +4 dgt)
Resolución, escala 400 W : 100 m W
Tensión de prueba: todas esc.: 0.4V
Protección sobrecarga, todas las
escalas: 500 V CC o CA ef.
Prueba de diodos
Escala: 4 V
Resolución: 1 mV
Precisión: ±(1.0% lect +1 dgt)
Corriente de prueba: 0.6 mA
Indicación: Tensión directa de la
unión
Tensión circuito abierto: 3.2 V CC
Protecc. sobrec.: 500 V CC o CA ef.
Indicador de continuidad
Escala: 400 W
Umbral audible: <40 W
Tiempo de respuesta: aprox. 100 ms
Tensión, circuito abierto: 400 mV CC
Protecc. sobrec.: 500 V CC o CA ef.
Capacidad
Escalas: 4-400 nF, 4-40 µF
Precisión, escala 4 nF: ±(1.0% lect
+40 dgt); 40 nF-4 µF: ±(1.0% lect
+4 dgt); 40 µF (<20 µF): ±(1.0%
lect +4 dgt); 40 µF (>20 µF): ±(5%
– 33 –
2015.qxd 7/21/00 5:41 PM Page 35
lect +8 dgt)
Frecuencia de prueba, todas esc. 1Hz
Nota: utilice el modo relativo para las
escalas de 4 nF y 40 nF
Resolución, escala 4 nF: 0.001 nF
Protecc. sobrec.: 500 V CC o CA ef.
Contador de frecuencia
Escalas: 100 Hz; 1, 10, 100, 400 KHz
Precisión, escala 100 Hz: ±(0.1% lect
+10 dgt)
1 y 10 KHz: ±(0.1% lect +4 dgt)
100 KHz: ±(0.1% lect +8 dgt)
400 KHz: ±(0.1% lect +20 dgt)
Resolución, escala 100 Hz: 0.01 Hz
Nivel de disparo, 100 Hz-100 KHz: 40
mV; 100 KHz-400 Kz: 400 mV
Protecc. sobrec.: 500 V CC o CA ef.
Nota: Para frecuencias por debajo de
1 Hz, el visualizador indica 00.00 Hz.
Accesorios opcionales
RF241 Sonda de RF 650 MHz
DL243C Juego de puntas de prueba
(básico)
DL248C Juego de puntas de prueba
(especial)
TL253A Convertidor de temperatura
(900°C/1652°F)
TL245 Puntas de prueba de seguridad
de repuesto
CT231A Pinza de corriente 150 A CA
CT232A Pinza de corriente 1000 A CA
CT234A Pinza de corriente 400 A CA
CT235 Pinza 1000 A CA/CC
CT236A Pinza 500 A CA (salida mV)
CT237 Pinza de corriente 200 A CA/CC
CT238 Pinza de corriente 20 A CA/CC
VC221A Estuche de vinilo acolchado.
Cabe el medidor con funda protectora
VC231 Estuche de vinilo acolchado.
Cabe el medidor sin funda protectora
DC205C Estuche de lujo con carcasa dura
DC207C Estuche de lujo grande con carcasa
dura y con espacio extra para accesorios
– 34 –
F SPECIFICATIONS
Spécifications Générales
Afficheur: LCD 3-3/4 digits, 4000
points (9999 points en mode
Fréquence), avec bargraphe à 40
segments, indicateurs de fonctions
et d’unités.
Indication de Polarité: Automatique,
positive impliquée, négative (–)
indiquée.
Adjustement du zéro: Automatique
Indication de dépassement de gamme:
Chiffre plus significatif “4”
clignotte. Autres chiffres indiqués
comme zéro.
Indication de pile déchargée: .
Changez la pile immédiatement
Raffraichissement: 2/sec, nominal;
1/sec en modes capacité et fré-
quence; 20/sec pour le bargraphe.
Temp. de Fonctionnement: 0 à 50°C,
0 à 70% Humidité Relative
Temp. de Stockage: -20°C à 60°C, 0 à
80% H.R., pile enlevée
Coefficient de Température: 0.1 x
(précison spécifiée)/°C (<18°C ou
>28°C°
Alimentation: Pile 9-Volt Standard,
NEDA 1604, JIS 006P, IEC 6F22
Coupure Automatique: après 30
minutes d’inactivité.
Autonomie (typique): 500 heures
alcaline, 250 heurs zinc-charbon
Dimensions (H x L x P): 175 x 90 x 34
mm
Poids (avec pile): 568 g
Accessoires: Cordons de test, fusible
de réserve, pile, manuel
d’utilisation
2015.qxd 7/21/00 5:41 PM Page 36
Sécurité: Conforme à EN61010-1
Cat.III 1000V; EMC: selon
EN50081-1, EN50082-2
EMC: Ce produit est con-
forme aux exigences des
directives suivantes de la
Communauté Européenne:
89/336/EEC (Compatibilité Elec-
tromagnétique) et 73/23/EEC
(Basse Tension), modifiée par
93/68/EEC (CE Marking).
Cependant, du bruit électrique ou
des champs électromagnétiques
intenses dans la proximité de
l’instrument peuvent influencer le
circuit de mesure. L’instrument
peut également être perturbé par
des signaux parasytes dans le
circuit mesuré. L’utilisateur doit
être vigilant et prendre des
précautions appropriées pour
éviter des résultats erronés quand
les mesures sont prises en
présence d’interférences électro-
magnétiques.
Spécifications Electriques
Précisions à 23°C ± 5°C, <75% H.R.
Tension Continue
Gammes: 400mV, 4, 40, 400, 1000V
Résolution, 100µV en gamme 400mV:
Précision, toutes gammes:
±(0.25%lect +2dgt)
Impédance d’entrée, gm 400mV:
>1000MW ; gm. 4V: 11M W; gms.
40V-1000V: 10MW
Protection de surcharge: 1100Vcc ou
crête ca. Transitoires de 6kV.
Tension Alternative (50-1000Hz).
Modèle 2015: Mesure efficace vraie. Les
précisions s’appliquent à toutes les
formes d’onde.
Gammes: 400mV, 4, 40, 400, 750V
Résolution: 100µV en gamme 400mV.
Factor crête: 3:1
Précision ±(%gamme +nbr digits):
Gamme 50-500Hz 500-1000Hz
400mV* ±(1.0% +5) non spécifié
40V-400V ±(0.8% +3) ±(1.0% +5)
Autres ±(0.8% +3) ±(1.2% +5)
Impédance d’entrée, gm. 400mV
>1000WM ; gm 4V: 11MW ; gms
40V-750V: 10MW
Protection de surcharge: 1100Vcc ou
crête ca. Transitoites de 6kV.
* Réponse Fréquence 50-100Hz
Courant Continu
Gammes: 4, 40, 400mA, 20A
Résolution, gamme 4mA: 1µA
Précision, gammes 4mA à 400mA:
±(0.8%lect + 2dgt)
gamme 20A: ±(1.2%lect ±4dgt)
Chute de tension: 500mV.
Gamme 400mA: 900mV
Protection de surcharge: Entrée mA:
fusible rapide 0.5A/250V (FP250 -
6.4 x 25.4 mm); pouvoir de
coupure 10kA
Entrée 20A: 10A en continu, 20A
pour 30 seconds. Fusible rapide
Littel Fuse KLK 20A ou Bussman
KTK 20A 10.3x38.1mm. 100kA
interrupt rating.Pouvoir de coupure:
100kA.
Courant Alternatif (50-1000Hz)
Modèle 2015: Mesure efficace vraie. Les
précisions s’appliquent à toutes les
formes d’onde.
Gammes: 4, 40, 400mA, 20A
Résolution, gamme 4mA: 1µA
Factor crête: 3:1
Précision, gammes 4mA à 400mA:
±(1.2%lect + 4dgt)
gamme 20A: ±(1.5%lect ±5dgt)
Chute de tension: voir courant cc.
Protection de surcharge: voir courant
cc.
– 35 –
2015.qxd 7/21/00 5:41 PM Page 37
Résistance
Gamme/Précision:
400
W
/±(0.8%lect. + 4dgt)
4k-400kW /±(0.7%lect + 2dgt);
4MW /±(1.0%lect + 4dgt);
40MW /±(1.5%lect + 4dgt)
Résolution, gamme 400W : 100mW
Tension de test, tts gammes: 0.4V
Protection de surcharge, toutes
gammes: 500VDC ou AC RMS
Test de Diodes
Gamme: 4V
Résolution: 1mV
Courant de test: 0.6mA
Affichage: Chute de tension en
direction passante
Précision: ±(1.0%lect +1dgt)
Tension en circuit ouvert: 3.2Vdc
Protection de surcharge: 500Vcc ou
ca effectif.
Indication de Continuité
Gamme: 400 Ohms
Seuil pour indication sonore: <40 W
Temps de réponse: approx. 100ms
Tension en circuit ouvert: 400mVcc
Protection de surcharge: 500Vcc ou
ca effectif.
Capacité
Gammes: 4, 40, 400nF, 4, 40µF
Précision, gamme 4nF: ±(1.0%lect +
40dgt); 40nF-4µF: ±(1.0%lect +
4dgt); 40µF (<20µF): ±(1.0%lect +
4dgt);
40µF (>20µF): ±(5.0%lect + 8dgt)
Fréquence test, tts gammes: 1Hz
Note: utilisez le mode relatif pour gammes
4nF et 40nF
Résolution, gamme 4nF: 0.001nF
Protection de surcharge: 500Vcc ou
ca effectif
Compteur de Fréquence
Gammes: 100Hz, 1, 10, 100, 400kHz
Précision, gm. 100Hz: ±(0.1%lect
+10dgt);
1 & 10kHz: ±(0.1%lect +4dgt);
100kHz: ±(0.1%lect +8dgt);
400kHz: ±(0.1%lect +20dgt)
Résolution, gamme 100Hz: 0.01Hz
Niveau de déclenchement - 100Hz-
100kHz: 40mV; 100kHz-400kHz:
400mV
Protection de surcharge: 500Vcc ou
ca effectif
Note: Pour fréquences <1Hz, affichage de
00.00Hz.
Accessoires en Option
RF241 Sonde RF 650MHz
DL243C Set de cordons standards
DL248C Set de cordons deluxe
TL253A Convertisseur de Températ.
(900°C/1652°F)
TL245 Cordons de remplacement
CT231A Pince de courant 150A ca
CT232A Pince de courant 1000A ca
CT234A Pince de courant 400A ca
CT235 Pince de courant 1000A ca/cc
CT236A Pince de courant 500A ca (sortie
mV)
CT237 Pince de courant 200A ca/cc
CT238 Pince de courant 20A ca/cc
VC221A Sacoche vynil rembourrée. Pour
instrument avec holster.
VC231 Sacoche vynil rembourrée. Pour
instrument sansholster.
DC205C Etui rigide de luxe
DC207C Large étui rigide de luxe avec
espace de rangement pour les
accessoires
– 36 –
2015.qxd 7/21/00 5:41 PM Page 38
TROUBLESHOOTING and MAINTENANCE
If there appears to be a malfunction during the operation of the meter, the
following steps should be performed in order to isolate the cause of the
problem: Check the battery. Review the operating instructions for
possible mistakes in operating procedure. Inspect and test the Test
Probes for a broken or intermittent connection. Inspect and test the
fuse. See Fuse Replacement.
Except for the replacement of the battery or fuse, repair of the multimeter
should be performed only by a Factory Authorized Service Center or by
other qualified instrument service personnel.
The front panel and case can be cleaned with a mild solution of detergent
and water. Apply sparingly with a soft cloth and allow to dry completely
before using. Do not use aromatic hydrocarbons or chlorinated solvents
for cleaning.
D • Fehlersuche und Unterhalt
Prüfen Sie zuerst folgende Fehlerquellen: Meßkabel (Brüche), Anschluß,
Zustand von Batterie und Sicherung, richtiger Meßvorgang, Eingangs- und
Bereichsgrenzen, usw.
Mit Ausnahme des Batterie- und Sicherungswechsels soll jede Reparatur
nur durch eine durch Wavetek anerkannte Servicestelle durchgeführt
werden.
Das Gerät kann mit einer milden Seifenlösung gereinigt werden. Sparsam
auftragen und vor Gebrauch gut trocknen lassen. Keine Lösungsmittel zum
Reinigen verwenden.
E • Mantenimiento y reparación
Para identificar la causa del problema: Compruebe la pila; revise las
instrucciones de uso; inspeccione las puntas de prueba por si hay una
conexión rota o intermitente; inspeccione la pila y el fusible.
Excepto la sustitución de la pila o el fusible, cualquier trabajo de reparación
del multímetro debe hacerse exclusivamente por personal técnico
cualificado para este tipo de reparaciones.
Para limpiar la carcasa puede utilizarse una solución suave de agua y
detergente.
F • Maintenance et Réparation
Avant d’expédier votre multimètre pour réparation, vérifiez les cordons de
mesure (rupture), pile et fusible, connections, procédure de mesure, limites
d’entrée et de calibres, etc.
– 37 –
2015.qxd 7/21/00 5:41 PM Page 39
Excepté pour le remplacement de la pile et du fusible, toute réparation doit
être effectuée uniquement par un Centre de Services agrée par Wavetek.
Vous pouvez nettoyer le boîtier avec un détergent doux. Appliquez
parcimonieusement et laissez sécher complètement avant utilisation. Ne pas
utiliser de solvents.
BATTERY / FUSE REPLACEMENT (Fig. 7)
Warning:
To prevent electrical shock hazard, turn off the multimeter and
any device or circuit under test and disconnect the test leads before
removing the rear cover.
– 38 –
Fig. 7
Input Fuse Type Dimensions Interrupt rating
Eingang Sicherung Typ Abmessungen Trennvermögen
Entrada Fusible Tipo Dimensiones
Niv. de interrup.
Entrée Fusible Type Dimensions Pouv. coupure
mA 0.5A/250V fast FP250 6.4x25.4mm 10kA
blow ceramic fuse
1)
20A 20A/600V fast Littel Fuse
blow ceramic fuse
1)
KLK 20A 10.3x38.1mm 100kA
Bussman KTK
1)
flinke Keramik-Sicherung // Fusible cerámico rápido // Fusible céramique rapide
2015.qxd 7/21/00 5:41 PM Page 40
After disconnecting test leads and turning off the multimeter, remove
the rear case by unscrewing the three screws that secure it to the front
case. a - mA Input Jack Fuse: Remove the blown fuse (6.4 x
25.4mm) from the fuse holder. Replace with a 0.5A/250V quick acting
fuse. Wavetek replacement fuse part number is FP 250 (a pack of four
fuses). b - 20A Input Fuse: Remove the 20A fuse (10.3x38.1mm)
from the 20 Ampere fuse holder. Replace with Wavetek FP425, Littel Fuse
KLK-20 or Bussman KTK-20. c - Battery replacement: Remove
battery and replace with NEDA type 1604, JIS006P, IEC 6F22 or
equivalent 9-volt alkaline battery. Reassemble the instrument.
Warning:
Use only an equivalent fuse to the one specified. Use of an
incorrect fuse could result in serious injury or even death.
Warning:
Failure to turn off the multimeter before installing the battery
could result in damage to the instrument and to the battery if the battery is
connected incorrectly to the multimeter.
D • Batterie/Sicherungsaustausch (siehe Fig 7)
Warnung: Um elektrischen Schock zu vermeiden, Gerät vor Öffnen
abschalten und Meßkabel entfernen
Geräterückseite losschrauben und abheben. a - mA Eingangs-
sicherung: Defekte 6.4 x 25.4 mm Sicherung entfernen und mit
0.5A/250V flinker Sicherung ersetzen (Teil Nummer FP250; 4er Packung).
b - 20A Eingangssicherung: Defekte 10.3 x 38.1 mm Sicherung
entfernen und mit gleichwertiger 20A flinker Keramik Sicherung ersetzen
(Wavetek FP425, Littel Fuse KLK-20 oder Bussman KTK-20). c -
Batterieaustausch: Batterie entfernen und durch gleichwertige 9V
Batterie (NEDA 1604, JIS006P, IEC 6F22) ersetzen. Geräterückseite
wieder anbringen und festschrauben.
Warnungen: Nur gleichwertige Sicherungen verwenden. Nicht-Abschalten
des Gerätes bei Batterieaustausch kann Batterie und Gerät zerstören.
E • Sustitución de la Pila y los Fusibles (vea fig. 7)
Advertencia: Para evitar el peligro de descarga eléctrica, apague el
multímetro y desconecte las puntas de prueba antes de abrir la tapa
posterior. Quite los tres tornillos y levante la tapa posterior. a Fusible
de la entrada de mA: retire el fusible abierto (6.4 x 25.4 mm) y
sustitúyalo por uno de actuación rápida y 0.5A/250V (ref. FP250, paquete
con 4). b Fusible de la entrada de 20 A: retire el fusible abierto
(10.3 x 38.1 mm) y sustitúyalo por otro cerámico, de actuación rápida y
20A (Wavetek FP425, Littel Fuse KLK-20 or Bussman KTK-20). c
Sustitución de la pila: Retire la pila usada y cámbiela por otra de 9V de
– 39 –
2015.qxd 7/21/00 5:41 PM Page 41
tipo NEDA 1604, JIS006P, IEC 6F22 o equivalente, preferiblemente alcalina.
Vuelva a cerrar el instrumento.
Advertencia: Utilice solamente fusibles equivalentes a los especificados.
Advertencia: Si no apaga el multímetro antes de cambiar la pila, podría
causar daños al instrumento y a la pila
F • Remplacement Pile et Batterie (voir fig 7)
Avertissement: Pour éviter des chocs électriques et/ou des dégats à
l’appareil, éteignez celui-ci et enlevez les cordons avant de l’ouvrir.
Dévissez le boîtier arrière et enlevez-le a - Fusible d’entrée mA:
Enlevez le fusible défectueux 6.4 x 25.4 mm mm et remplacer le par un
nouveau fusible rapide 0.5A/250V (Référence FP250; boîte de quatre). b
- Fusible d’entrée 20A: Enlevez le fusible défectueux 10.3 x 38.1 mm
Sicherung et remplacer le par un nouveau fusible céramique rapide 20A
(Wavetek FP425, Littel Fuse KLK-20 ou Bussman KTK-20). c -
Remplacement de la pile: Enlevez la pile, et remplacez la par une pile
équivalente 9V (NEDA 1604, JIS006P, IEC 6F22). Réassemblez
l’appareil.
Avertissements: Ne remplacez les fusibles que par des fusibles
équivalents. Ne pas éteindre l’appareil pour le remplacement des piles peut
endommager la plie et l’appareil.
REPAIR
Read the warranty located at the front of this manual before requesting
warranty or non-warranty repairs. For warranty repairs, any multimeter
claimed to be defective can be returned to any Wavetek Meterman
authorized distributor or to a Wavetek Meterman Service Center for an
over-the-counter exchange for the same or like product. Non-warranty
repairs should be sent to a Wavetek Meterman Service Center. Please call
Wavetek Meterman or enquire at your point of purchase for the nearest
location and current repair rates. All multimeters returned for warranty or
non-warranty repair or for calibration should be accompanied by the
following information or items: company name, customer's name,
address, telephone number, proof of purchase (warranty repairs), a brief
description of the problem or the service requested, and the appropriate
service charge (for non-warranty repairs). Please include the test leads
with the meter. Service charges should be remitted in the form of a check,
a money order, credit card with expiration date, or a purchase order made
payable to Wavetek Meterman or to the specific service center. For
minimum turn-around time on out-of-warranty repairs please phone in
advance for service charge rates. The multimeter should be shipped with
– 40 –
2015.qxd 7/21/00 5:41 PM Page 42
transportation charges prepaid to one of the following addresses or to a
service center:
iinn UU..SS..AA.. iinn CCaannaaddaa iinn EEuurrooppee
Wavetek Meterman Wavetek Meterman Wavetek Meterman
1420 75th Street SW 400 Britannia Rd. E.Unit #1 52 Hurricane Way
Everett, WA 98203 Mississauga, ON L4Z 1X9 Norwich, NR6 6JB, U.K.
Tel: 1-877-596-2680 Tel: (905) 890-7600 Tel:int + 44-1603-404824
Fax: 425-446-6390 Fax: (905) 890-6866 Fax:int + 44-1603-48240
9
The instrument will be returned with the transportation charges paid by
Wavetek Meterman.
D • Reparatur
Lesen Sie die Gewährl ei stung bevor Sie eine Reparat ur unter oder
außerhalb Gewährleistung anfragen. Unter Gewährleistung bringen Sie bitte
das def ekte Gerät zu ei ner aner kannten Wavetek Meterman Verkaufsstelle
oder Servicestelle füreinen direkten Umtausch. Außerhalb Gewährlei st ung
senden Sie das Gerät zu einer Wavetek Meterman anerkannten Servicestelle.
Bitte informieren Sie sich bei Wavetek Meterman oder ihrem Fachhändler
nach der dichtst beigelegen Adresse und nach aktuellen Reparaturgebühren.
Bitte senden Sie folgende Informationen und Dokumente mit: Firmenname,
Kundenname, Adresse, Telefoonnummer, Kaufnachweis (für Reparaturen
unter Gewährl eistung), ei ne kurz e Beschr eibung der gewünschten
Handlung, und di e gefordert e Bezahl ung ( Eingr iffe außerhal b der
Gewährleistung). Bitte auch Testkabel beifügen. Bezahlungen in Form eines
Checks, Bezahlungsformulieren, Kredietkarte mit Verf alldatum, usw. bitte in
Namen der Servicestelle aufstellen. Bitte Multimeter (Frei) senden an:
iinn UU..SS..AA.. iinn CCaannaaddaa iinn EEuurrooppaa
Wavetek Meterman Wavetek Meterman Wavetek Meterman
1420 75th Street SW 400 Britannia Rd. E.Unit #1 52 Hurricane Way
Everett, WA 98203 Mississauga, ON L4Z 1X9 Norwich, NR6 6JB, U.K.
Tel: 1-877-596-2680 Tel: (905) 890-7600 Tel:int + 44-1603-404824
Fax: 425-446-6390 Fax: (905) 890-6866 Fax:int + 44-1603-482409
oder an die Ihnen mitgetei lte Adresse. Multimeter wird (Frei) zurück
geschickt
E • Reparación
Lea las condiciones de garantía, al principio de este manual, antes de
solicitar cualquier reparación dentro o fuera de garantía. Si la reparación es
en garantía, puede llevar el multímetro defectuoso a cualquier Distribuidor
– 41 –
2015.qxd 7/21/00 5:41 PM Page 43
Autorizado o Centro de Servicio de Wavetek Meterman, donde le cambiarán
en mano el producto por otro igual o similar. Para reparaciones fuera de
garantía deberá enviar el multímetro a un Centro de Servicio de Wavetek
Meterman. En Wavetek Meterman, o en su Distribuidor o punto de venta, le
indicarán el Centro de Servicio más próximo y las tarifas de reparación
vigentes. La documentación que acompañe a todo multímetro enviado para
reparación debe incluir los siguientes datos: nombre de la empresa, persona
de contacto, dirección, número de teléfono, prueba de compra (para
reparaciones en garantía), una breve descripción del problema o el servicio
requerido y, en caso de reparaciones fuera de garantía, si desea presupuesto
previo. Por favor envíe las puntas de prueba con el multímetro. El importe de
la reparación se enviará en forma de cheque, tarjeta de crédito con fecha de
expiración u orden de pago a favor de Wavetek Meterman o del Centro de
Servicio específico. El multímetro se enviará a portes pagados a una de las
siguientes direcciones, o al Centro de Servicio que le hayan indicado:
eenn EEEE.. UUUU.. eenn CCaannaaddáá eenn EEuurrooppaa
Wavetek Meterman Wavetek Meterman Wavetek Meterman
1420 75th Street SW 400 Britannia Rd. E.Unit #1 52 Hurricane Way
Everett, WA 98203 Mississauga, ON L4Z 1X9 Norwich, NR6 6JB, U.K.
Tel: 1-877-596-2680 Tel: (905) 890-7600 Tel:int + 44-1603-404824
Fax: 425-446-6390 Fax: (905) 890-6866 Fax:int + 44-1603-482409
Wavetek Meterman devolverá el multímetro reparado a portes pagados.
F • Réparation
Lisez la garantie au début de ce manuel avant de demander une réparation
sous garantie ou hors gar antie. Pour une réparation sous garantie ,
adressez-vous à votre revendeur Wavetek Meterman ou à un centre de
services agréé par Wavetek Meterman pour un échange direct. Pour une
réparation hors garantie, envoyez votre multimètre à un Centre de Services
agrée par Wavetek Meterman . Téléphonez à Wavetek Meterman ou
demandez à votre revendeur pour l'adresse la plus proche. Pour les
réparations hors garantie, demandez dabord les tarifs. Joignez les
informations et documents suivants: nom de sociètè, nom du client,
adresse, numéro de téléphone, preuve d'achat (pour réparations sous
garantie), une brève description de l'intervention souhaitée et le payement
(pour réparations hors garantie). Ajoutez également les cordons de test. Le
payement, sous forme de chèque, virement, carte de crédit avec date
d'expiration, etc. doit êtr e éta bli au nom du Centre de Servic es. Le
multimètre doit être envoyé port payé à:
– 42 –
2015.qxd 7/21/00 5:41 PM Page 44
eenn UU..SS..AA.. eenn CCaannaaddaa eenn EEuurrooppaa
Wavetek Meterman Wavetek Meterman Wavetek Meterman
1420 75th Street SW 400 Britannia Rd. E.Unit #1 52 Hurricane Way
Everett, WA 98203 Mississauga, ON L4Z 1X9 Norwich, NR6 6JB, U.K.
Tel: 1-877-596-2680 Tel: (905) 890-7600 Tel:int + 44-1603-404824
Fax: 425-446-6390 Fax: (905) 890-6866 Fax:int + 44-1603-482409
ou à l'adresse communiquée. Le multimètre vous sera renvoyé port payé.
– 43 –
2015.qxd 7/21/00 5:41 PM Page 45
®
T
Manual Revision Date 07/00
Manual Part Number 1566149
Information contained in this
manual is proprietary to Wavetek
Meterman and is provided solely
for instrument operation and
maintenance. The information in
this document may not be
duplicated in any manner without
the prior approval in writing from
Wavetek. Meterman.
Specifications subject to change.
WAVETEK is a trademark of Wavetek
Wandel Goltermann.
© Wavetek Meterman, 2000
U.S. Service Center
Wavetek Meterman
1420 75th Street SW
Everett, WA 98203
Tel: (877) 596-2680
Fax: 425-446-6390
Canadian Service Center
Wavetek Meterman
400 Britannia Rd. E.Unit #1
Mississauga, ON L4Z 1X9
Tel: (905) 890-7600
Fax: (905) 890-6866
European Distribution Center
Wavetek Meterman
52 Hurricane Way
Norwich, NR6 6JB, England
Tel: (44) 1603-404-824
Fax: (44) 1603-482-409
2015.qxd 7/21/00 5:41 PM Page 46

Transcripción de documentos

2015.qxd 7/21/00 5:40 PM Page 1 ® TM 2015.qxd 7/21/00 5:40 PM Page 2 WARRANTY The 2015 Digital Multimeter is warranted against any defects of material or workmanship within a period of one (1) year following the date of purchase of the multimeter by the original purchaser or original user. Any multimeter claimed to be defective during the warranty period should be returned with proof of purchase to an authorized Wavetek Meterman Service Center or to the local Wavetek Meterman dealer or distributor where your multimeter was purchased. See maintenance section for details. Any implied warranties arising out of the sale of a Wavetek Meterman multimeter, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited in duration to the above stated one (1) year period. Wavetek Meterman shall not be liable for loss of use of the multimeter or other incidental or consequential damages, expenses, or economical loss or for any claim or claims for such damage, expenses or economical loss. Some states do not allow limitations on how long implied warranties last or the exclusion or limitation of incidental or consequential damages, so the above limitations or exclusions may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state. D GEWÄHRLEISTUNG Die Digitale Multimeter Modelle 2015 ist ab Kaufdatum für ein (1) Jahr gegen Material- und Herstellungsfehler gewährleistet. Siehe Kapitel "Reparatur" für Einzelheiten. Implizierte Schadeforderungen sind auch auf ein Jahr beschränkt. Wavetek Meterman ist nicht ansprechbar für Gebrauchsverluß oder Folgeschäden, Ausgaben, Gewinnverluß, usw. E GARANTIA Este Multímetro Digital Modelo 2015 está garantizado contra cualquier defecto de material o de mano de obra durante un periodo de un (1) año contado a partir de la fecha de adquisición. En la sección de "Reparación" se explican los detalles relativos a reparaciones en garantía. Cualquier otra garantía implícita está también limitada al periodo citado de un (1) año. Wavetek Meterman no se hará responsable de pérdidas de uso del multímetro, ni de ningún otro daño accidental o consecuencial, gastos o pérdidas económicas, en ninguna reclamación a que pudiera haber lugar por dichos daños, gastos o pérdidas económicas. F GARANTIE Le multimètre digitaux, Modèle 2015 est garanti pour un (1) an à partir de la date d'achat contre les défauts de matériaux et de fabrication. Voir chapitre "Réparation" pour plus de détails. Toute garantie impliquée est également limitée à un an. Wavetek Meterman ne peut être tenu responsable pour perte d'utilisation ou autres préjudices indirects, frais, perte de bénéfice, etc. 2015.qxd 7/21/00 CONTENTS 5:40 PM Page 3 Safety Information ............................................... 2 Instrument Familiarization.................................... 5 Measurement Procedures ................................... 8 Specifications ................................................... 28 Troubleshooting and Maintenance .................... 37 D • Inhalt Sicherheitsinformationen .................................... 2 Vorstellung des Gerätes ....................................... 5 Meßprozeduren ................................................... 9 Spezifikationen .................................................. 30 Fehlersuche und Unterhalt ................................ 37 E • Contenido Información de seguridad ................................... 3 Familiarización con el instrumento ...................... 5 Procedimientos de medida .................................. 9 Especificaciones ............................................... 32 Localización de avería y mantenimiento............. 37 F • Contenu Informations de Sécurité ..................................... 3 Présentation de l’Appareil ................................... 5 Procédures de Mesure ...................................... 10 Spécifications ................................................... 34 Dépannage et Maintenance ................................ 37 EXPLANATION OF SYMBOLS D • Erklärung der Symbole = E • Significado de los símbolos = F • Explication des Symboles Direct current •D• Gleichstrom •E• Corriente continua •F• Courant continu. Alternating current •D• Wechselstrom •E• Corriente alterna •F• Courant alternatif. Ground connection •D• Erdanschluß •E• Conexión a tierra •F• Connection de terre. Attention! Refer to Operating Instructions •D• Achtung! Bitte Anleitung lesen •E• ¡Atención! Consulte las Instrucciones de Uso •F• Attention! Consultez le manuel. Dangerous voltage may be present at terminals •D• Eine gefährliche Spannung kann an den Eingängen anliegen •E• Puede haber tensión peligrosa en los terminales •F• Une tension dangereuse peut être présente aux entrées. This instrument has double insulation •D• Dieses Gerät ist doppelt geisoliert •E• Este instrumento tiene doble aislamiento •F• Cet appareil est prévu d’une double isolation. –1– 2015.qxd 7/21/00 5:40 PM Page 4 WARNINGS AND PRECAUTIONS n This instrument is EN61010-1 certified for Installation Category III. It is recommended for use in distribution level and fixed installations, as well as lesser installations, and not for primary supply lines, overhead lines and cable systems. n All inputs are protected against continuous overload conditions up to the limits of each function's stated input protection (see Specifications). Never exceed these limits nor the ratings marked on the instrument itself. n Voltages applied to COM terminal may be present at all other terminals n Exercise extreme caution when: measuring voltage >20V, current >10mA, AC power line with inductive loads, AC power line during electrical storms. High voltages can be lethal and high voltage transients may occur at any time. n Operator injury or damage to the multimeter may occur during current measurements if the fuse blows in a circuit with open circuit voltage >600V (250 V in mA input). n Always inspect your DMM, test leads and accessories for sign of damage or abnormality before every use. If abnormal condition exist (broken or damaged test leads, cracked case, display not reading, etc.), do not use. n When testing for voltage or current, make sure these ranges function correctly. Take a reading of a known voltage or current first. n Never ground yourself when taking measurements. Do not touch exposed metal pipes, outlets, fixtures, etc., which might be at ground potential. Keep your body isolated from ground and never touch exposed wiring, connections, test probe tips, or any live circuit conductors. n Always measure current in series with the load – NEVER connect the multimeter ACROSS a voltage source. Check fuse first. n Never replace a fuse with one of a different rating. n Do not operate instrument in an explosive atmosphere (flammable gases, fumes, vapor, dust.) n Do not use this or any piece of test equipment without proper training n CRT SERVICE SAFETY REMINDER : A potential danger exists when measuring voltages in the horizontal output and damper stages of CRT equipment. (High voltage transients greater than 6,000 V). Refer to your CRT service manual for proper servicing instructions. D • Warnungen und Vorsichtsmaßnahmen n Dieses Gerät ist EN61010-1 zertifiziert für Installationsklasse III. Anwendung ist empfohlen auf Verteilerebene und festen Anlagen sowie untergeordneten Systemen, jedoch nicht für Starkstromnetze und Hochspannungsanlagen. n Überschreiten Sie nie die kontinuierlichen Überlastgrenzen der verschiedenen Meßfunktionen (siehe Spezifikationen) –2– 2015.qxd 7/21/00 5:40 PM Page 5 oder andere Grenzen welche auf dem Gerät markiert sind. n Spannungen welche am COM Eingang anliegen können sich auch auf andere Eingänge übertragen nVorsicht beim Messen von Spannungen >20V // Strömen >10mA // Netzstrom/-spannung bei induktiever Last oder bei Gewittern // Strom, wenn die Sicherung durchbrennt in einem Schaltkreis mit Leerlaufspannung >600V (>250V beim mA Eingang) // beim Messen an Bildröhrgeräten (hohe Spannungsspitzen) n Unsersuchen Sie Gerät, Meßkabel, Verbinder, usw. vor jeder Messung. Beschädigte Teile nicht verwenden n Meßspitzen und Stromkreis während der Messung nicht berühren • Sich selbst isolieren ! n Bei Strommessung, Multimeter immer in Serie mit Schaltkreis verbinden – Nie in parallel mit Spannungsquelle. n Sicherung immer mit gleichwertiger ersetzen. n Gerät nicht in explosiver Umgebung verwenden. E • Advertencias y Precauciones n Este instrumento está homologado según EN61010-1 para la Categoría de Instalación III. Su uso está recomendado en el nivel de distribución y en instalaciones fijas, así como en instalaciones menores, pero no en líneas principales de suministro, líneas aéreas ni sistemas de cable. n No supere nunca los límites de entrada para las diferentes funciones (vea Especificaciones), ni los límites marcados en el instrumento. n Tensiones presentes en COM puedan estar presentes en otras entradas. n Tenga especial cuidado al: medir tensión >20 V // corriente >10 mA // tensión de red de CA con cargas inductivas // tensión de red de CA durante tormentas eléctricas // corriente, si salta el fusible en un circuito con tensión de circuito abierto >600 V (250 V en la entrada de mA) // trabajar con pantallas TRC n Inspeccione siempre el multímetro, las puntas de prueba, los conectores y los accesorios antes de cada uso. No utilice ningún componente que esté dañado. n No se ponga Ud. a tierra cuando esté tomando medidas, y no toque nunca partes expuestas de los circuitos. n Al medir corriente, conecte siempre el multímetro EN SERIE con la carga - NUNCA EN PARALELO con una fuente de tensión. n Nunca sustituya un fusible con otro que no tenga las mismas especificaciones. n No utilice el instrumento en ambientes potencialmente explosivos. F • Avertissements et Précautions n Cet instrument est certifié EN61010-1 catégorie d’installation III. Son utilisation est recommandée pour le niveau de distribution de réseau, les installations fixes et systèmes subordonnés, et non pour les installations de puissance et lignes de transmission et câblages à haute tension. n N’excédez jamais les limites de surcharge continues (voir spécifications) ou d’autres limites marquées sur l’appareil. n Des tensions appliquées à l’entrée COM peuvent être également présentes aux autres bornes d’entrée. n Soyez très prudent quand vous mesurez des tensions >20V ou des –3– 2015.qxd 7/21/00 5:40 PM Page 6 courants >10mA // tension ou courant de secteur avec charge inductive ou par temps de tempête // du courant quand le fusible saute dans un circuit avec tension en circuit ouvert de >600 volts (>250V pour l’entrée mA) // dans des appareils à tube cathodique (transitoires à haute tension) n Inspectez appareil, câbles, connecteurs avant chaque mesure. N’utilisez pas des pièces endommagées n Ne touchez pas les pointes de touche ou le 3 3/4 digit LCD with Unit, Status and Function Indicators 3 3/4 Digit LCD ; Dezimalpunkt, Funktions- und Einheitsanzeigen LCD de 3 3/4 dígitos con indicadores de unidades, estado y función Low Battery Batterie entladen Pila baja Pile déchargée LCD 3 3/4 digits ; point décimal, annonciateurs de fonctions et d’unités Special Function Buttons Spezielle Funktionstasten Teclas de función especial Boutons de fonctions spéciales 40-Segment Bargraph 40-Segment Bargraf Barra analógica Bargraphe 40-segments High input for voltage, resistance, frequency, capacitance Function/Range/Off Selector Funktion-/Bereich/Aus Schalter Selector de Función/Escala/Off Sélecteur fonctions/ calibres/marche-arrêt 20 A Input 20 A Eingang Entrada 20A Entrée 20 A 400mA Input 400mA Eingang Entrada 400mA Entrée 400mA V HzC Eingang. Hoch für Spannung Widerstand, Frequenz, Kapazität Entrada “alta” para tensión, resistencia y capacidad Entrée V HzC. Haut pour tension, résistance, fréquence, capacité COM Input – common or low input for all measurements COM Eingang – Referenzpunkt für alle Messungen COM- entrada común o “baja” para todas las medidas Entrée COM – commun ou bas pour toutes mesures –4– 2015.qxd 7/21/00 5:40 PM Page 7 circuit pendant les mesures. Isolez-vous ! n Pour la mesure de courant, connectez l’appareil en série avec le circuit – JAMAIS en parallèle avec une source de tension. n Remplacez les fusibles toujours par des fusibles équivalents n N’utilisez pas cet appareil dans des atmosphères explosives. PREPARATION FOR USE – UNPACKING Your shipping carton should include the multimeter, a holster, one test lead set (one black, one red), one 9V battery (installed), a warranty card and this manual. If any of the items are damaged or missing, immediately return the complete package to the place of purchase for an exchange. D • Gebrauchsvorbereitung - Auspacken Die Verpackung sollte enthalten: ein Multimeter, ein Meßkabelsatz (ein schwarz, ein rot), eine 9V Batterie (im Gerät), eine Garantiekarte und diese Anleitung. Wenn ein Teil fehlt oder beschädigt ist, bitte bei der Verkaufsstelle umtauschen. E • Preparación del multímetro para su uso - Desembalaje El embalaje debe contener: el multímetro, una funda protectora, un juego de puntas (una negra y otra roja), una pila de 9 V (instalada), una tarjeta de garantía y este manual. Si falta algún componente u observa daños, devuelva el conjunto al lugar donde lo adquirió para que se lo cambien. E • Préparation pour l’Emploi - Déballage Votre emballage doit contenir: un multimètre, un jeu de câbles de mesure (un rouge, un noir), une pile 9V (installée), une carte de garantie et ce manuel. Si une pièce manque ou est endommagée, retournez à votre point de vente pur un échange. PROTECTIVE HOLSTER The holster/tilt stand protects the meter from accidental falls and provides greater ease of use. The tilt stand has two positions, and a velcro-strap allows to hang the meter from a fixture. Both test lead probes can be attached to the holster for storage. One probe can be attached for measurement, holding the meter with probe in one hand and the second probe in the other hand. –5– 2015.qxd 7/21/00 5:40 PM Page 8 D • Schutz-Holster Das Schutzholster beschirmt das Gerät vor Stürzen und Stößen. Der Kippständer hat zwei Positionen, und mit der Schlaufe können Sie das Gerät aufhängen. Die Meßsonden können am Holster befestigt werden. E • Funda Protectora La funda con pie integrado protege el medidor en caso de caídas accidentales. El pie inclinado tiene dos posiciones, y una banda con Velcro permite colgar el multímetro. Las puntas de prueba pueden fijarse a la funda. F • Holster de Protection Le holster protège l’appareil contre des chutes accidentelles et des chocs. Il est muni d’une béquille à deux positions et d’une boucle de suspension. Les sondes des cordons peuvent être attachées au holster. OVERLOAD INDICATION Range Overload is indicated by a flashing “4” (MSD) and three non-flashing zeros (plus a beep tone for voltage, capacitance and current). Take immediate steps to remove the cause of the overload. Select the next higher range until a value is displayed. If overload still exists in the highest range, remove test leads from the measurement setup as it is beyond the range of the meter. Overload indication is normal in the OHMS and continuity ranges when the leads are not connected to anything or when the measured value is higher than the selected resistance range. –6– 2015.qxd 7/21/00 5:40 PM Page 9 D • Überlastanzeige Bei Bereichsüberschreitung blinkt die höchste Ziffer “4”; andere Digits werden als “0” angezeigt. In Spannungs-, Kapazitäts- und Strombereichen ertönt gleichzeitig ein Summton. Ursache der Bereichsüberschreitung sofort entfernen (Höheren Bereich wählen oder Messung unterbrechen). Überlastanzeige ist normal in Widerstands- und Durchgangsbereichen bei offener Schaltung oder zu hohem Widerstand. E • Indicación de Sobrecarga La sobrecarga de escala se indica mediante el parpadeo del dígito más significativo (“4”) y tres ceros fijos (más un bip o tono audible en tensión, capacidad y corriente). Para eliminar la causa de la sobrecarga, seleccione una escala más ata o interrumpa la medida. La indicación de sobrecarga es normal en medida de OHMS y prueba de continuidad, cuando el circuito está abierto o la resistencia es demasiado alta. F • Indication de Surcharge Un dépasssement de calibre est indiqué par le clignottement du chiffre le plus significatif “4”; les autres chiffres étant “0”. Cet affichage est accompagné d’un signal sonore dans les gammes de tension, de capacité et de courant. Ecartez immédiatement la cause du dépassement – choisissez une gamme plus élevée ou interrompez la mesure. L’indication de dépassement est normale dans les gammes de résistance ou de continuité avec circuit ouvert ou résistance trop élevée. –7– 2015.qxd 7/21/00 5:40 PM Page 10 MEASURING PROCEDURES GENERAL: Turn instrument on by turning function/range switch away from OFF and selecting the parameter you want to measure. Six switch positions have dual functions (resistance/continuity, diode/capacitance, four AC/DC current ranges). How to select either is explained under the respective function. Ranging: This instrument is autoranging on all ranges except current and 200mV AC. It automatically selects the range that gives the best resolution for the measured value. “AUTO” in the display confirms that you are in Autoranging mode. You can tell which range (and function) you are in by the position of the decimal point and the measurement unit displayed. In the example shown you would be in the 40V AC range. Hold a range by pressing the RANGE button. Cycle through ranges by continuing to depress RANGE. Press and hold RANGE button for two seconds to return to Autoranging. Bargraph: The 40-segment bargraph indicates at which percentage of the range you presently measure. It updates 20 times per second, and, like an analog needle display, it is ideal for circuit adjustments. Display Hold: When you need to concentrate on your circuit, you can push the HOLD button to keep the reading on the display. “HLD” appears in the display. Push the HOLD button again to resume normal display of current value. (Automatic shut-down is disabled while “HLD” is displayed). General Measurement Procedures: Œ When connecting or disconnecting test leads to a circuit, always first turn off power to device or circuit being tested and discharge all capacitors.  Strictly observe the max input limits. Ž Do not change functions while test leads are connected to circuit. –8– 2015.qxd 7/21/00 5:40 PM Page 11 D • Meßprozeduren Allgemein: Gerät einschalten durch Funktionsschalter von OFF weg, nach einer gewünschten Meßfunktion zu drehen. Bereichswahl: Automatisch für alle Funktionen/Bereiche, ausgenommen Strom und 400mV AC. “AUTO” wird angezeigt. Aktuelle Funktion und Bereich werden durch Dezimalpunkt und Meßeinheit angezeigt. Sie können einen Bereich festhalten durch die RANGE Taste zu drücken. Sie können die verschiedenen Bereiche manuell durchlaufen durch wiederholt RANGE zu drücken. RANGE zwei Sekunden lang drücken um erneut automatische Bereichswahl zu wählen. Bargraf: Der 40-Segment Bargraf zeigt an an welcher Stelle des Bereiches (%) Sie messen. Erneuerung: 20 pro Sekunde. Anzeigenstop: HOLD Taste drücken um einen Meßwert auf der Anzeige festzuhalten. “HLD” wird angezeigt. HOLD erneut drücken um aktuelle Meßwerte anzuzeigen. Bei aktiver HOLD Funktion schaltet das Gerät nicht automatisch ab. Allgemeine Meßprozeduren: Œ Beim Verbinden/Trennen der Meßkabel mit dem/vom Schaltkreis, zuerst Schaltkreis abschalten und Kondensatoren entladen.  Eingangsgrenzen beachten. Ž Wahlschalter nicht drehen wenn Meßkabel mit dem Meßkreis verbunden sind. E • Procedimientos de medida EN GENERAL: Encienda el instrumento girando el selector de función/escala en otra posición que no sea OFF, seleccionando el parámetro que desee medir. Selección de escala: Este instrumento selecciona automáticamente la escala (auto-rango) en todas las medidas excepto corriente y 400mV AC. Aparece la indicación “AUTO”. La escala y la función seleccionadas en cada momento se indican mediante el punto decimal y la indicación de unidades. Para mantener fija una escala, pulse la tecla RANGE. Para recorrer todas las posibles escalas, pulse repetidamente RANGE. Para volver a AUTO, pulse y mantenga pulsada RANGE durante dos segundos. Barra analógica: La barra de 40 segmentos indica el porcentaje del fondo de escala en el que está midiendo en cada momento. La información se actualiza 20 veces por segundo. Congelación de la lectura: Pulse la tecla HOLD para mantener una lectura congelada en el visualizador. Aparece la indicación “HLD”. Pulse de nuevo HOLD para volver a la indicación de la lectura real. –9– 2015.qxd 7/21/00 5:40 PM Page 12 Procedimientos generales de medida: Œ Antes de conectar o desconectar las puntas de prueba a/de un circuito , apague siempre el dispositivo o circuito sometido a prueba y descargue todos los condensadores.  Observe estrictamente los límites máximos de entrada. Ž No cambie de función mientras tenga las puntas de prueba conectadas a un circuito. F • Procédures de Mesure Général: Mettez l’appareil sous tension en tournant le sélecteur de la position OFF vers la fonction souhaitée. Sélection de gammes: Automatique pour toutes les fonctions et gammes, sauf le courant et 400mV ca. “AUTO” est affiché. La fonction et la gamme en cours sont affichées par le point décimal et l’unité. Vous pouvez bloquer une gamme en presssant le bouton RANGE. Continuez à presser RANGE pour parcourir les diverses gammes. Pressez RANGE pendant deux secondes pour retourner à la sélectiion automatique. Bargraphe: Le bargraphe à 40-segments indique à quelle position de la gamme (%) vous mesurez. Il est raffraichi 20 fois par seconde. Maintien de lecture: Pressez la touche HOLD pour maintenir un affichage. “HLD” est affiché. Pressez HOLD à nouveau pour afficher les valeurs actuelles. La coupure automatique est désactivée pendant que HOLD est actif. Procédures Générales: Œ Avant de connecter/déconnecter les câbles de mesure à/d’un ciruit, coupez l’alimentation du circuit et déchargez les condensateurs.  Ne dépassez pas les limites d’entrée. Ž Ne tournez pas le sélecteur tandis que les cordons sont raccordés à un circuit. DC AND AC VOLTAGE MEASUREMENT (See Fig. 1) Œ Connect test leads as shown in figure 1.  Turn function switch to Vac ( ) or Vdc ( ). Ž Touch Probe tips across voltage source (in parallel with circuit).  Voltage value will appear on Digital Display along with the voltage polarity (DC – positive implied). Note: The 400mV AC range needs to be selected manually. See page 8 “RANGING” for procedure. Note for AC Measurements: Model 2015 is a True-RMS meter. It measures the true RMS value of distorted AC voltage or current signals. The Crest Factor handling capability is shown in table 1. The Crest Factor is the Peak Voltage divided by the RMS voltage. – 10 – 2015.qxd 7/21/00 5:40 PM Page 13 Fig. 1 Wave form, Crest Factor Wellenform, Crest Factor Forme d’onde, Facteur CrŒte Table 1 Additional correction from 1.5 to 5.0 Zus tzliche Korrektur von 1.5 bis 5.0 Correction supplØm. de 1.5 5.0 Input RMS, Percentage of Full-Scale Effektiv-Wert Eingang, % vom Endbereich EntrØe effective, % pleine Øchelle – 11 – 2015.qxd 7/21/00 5:40 PM Page 14 D • Gleich- und Wechselspannungsmessung (siehe Fig.1) Œ Meßkabel wie in Figur 1 gezeigt verbinden.  Funktionsschalter auf VDC ( ) oder VAC ( ) stellen. Ž Meßspitzen mit Meßkreis verbinden – parallel zur Spannungsquelle.  Meßwert ablesen (automatische Polaritätsanzeige bei DC Messungen). Anmerkung: Der 400mV AC Bereich muß manuell gewählt werden. Siehe Seite 9 “Bereichswahl” für Prozedur. Anmerkung für AC Messungen: Modell 2015 ist ein EchtEffektivwertmeßgerät: Es mißt den Effektivwert von verzerrten Spannungs- und Stromsignalen. Crest Faktor Grenzen sind in Tabelle 1 angezeigt. Der Crest Factor ist die Spitzenspannung dividiert durch die Effektivwertspannung. E • Medidas de tensión CC y CA (vea Fig. 1) Œ Conecte las puntas de prueba como se muestra en la Figura 1.  Ponga el selector de función en ACV o DCV. Ž Toque con las puntas de prueba los puntos de tensión (en paralelo con el circuito).  Aparece el valor de la tensión en el visualizador digital, junto con la polaridad en l caso de DCV. Nota: La escala de 400mV AC debe seleccionarse manualmente. Consulte el procedimiento en página 9, “Selección de escalas”. Nota para medidas de CA: El Modelo 2015 es un medidor de verdadero valor eficaz (TRMS). Mide el verdadero valor eficaz de señales distorsionadas de tensión o corriente CA. En la tabla 1 se muestra su capacidad de proceso del factor de cresta. El factor de cresta es la tensión máxima (pico) dividida por la tensión eficaz (RMS). F • Mesure de Tensions CC et CA (voir fig. 1) Œ Connectez les câbles comme indiqué en figure 1.  Placez le sélecteur sur VDC ( ) ou VAC ( ). Ž Connectez les pointes de touche au circuit – en parallèle avec la source de tension.  Lisez la valeur. Note: Le calibre 400mV ca doit être sélectionné manuellement. Voir page 10 “Changement de calibres” pour procédure. Notes pour mesure en courant alternatif: Le modèle 2015 est un appareil à mesure efficace vraie: Il mesure la valeur efficace de toutes les formes d’ondes, avec ou sans distortion. Les limites de facteur de crête sont indiquées dans le tableau 1. Le facteur de crête est la valeur crête divisée par la valeur efficace. – 12 – 2015.qxd 7/21/00 5:40 PM Page 15 DC AND AC CURRENT MEASUREMENT (See Fig. 2) Note: the current function is not autoranging. “AUTO” does not appear in the display, and “MANU” is displayed instead. The meter has four current ranges, which each have their own selector switch position. Œ Connect red test lead to the 400mA Input for current measurements up to 400mA or to the 20A input for current measurements to twenty amperes (20A for 30 seconds). Connect black test lead to COM Input Connector.  Set the Function Switch to the corresponding Current range – to A when the test lead is connected to the 20A input (mA ranges do not work with 20A input); to 4, 40 or 400mA when test lead is connected to the 400mA input (depending on how small a current you are going to measure). Ž By default, when you select a current function, your multimeter is in AC mode. “AC” is shown in the display. Push the yellow SHIFT button to select DC mode. “DC” is shown in the display (the SHIFT button toggles between the two modes).  Open circuit in which current is to be measured. Securely connect test leads in series with the load.  Turn on power to circuit being tested. ‘ Read current value on Digital Display. With 400mA input used, switch from 400mA to 40 or 4mA range if setting is too high (If you are in DC mode, the mutimeter falls back into AC mode each time you turn the selector switch – so you also each time need to press the yellow SHIFT button to return into DC mode). Note 1: Each current input has a protective fuse installed. If fuse does blow, see page 38 for replacement information. Note 2 – Incorrect Input Warning: The buzzer sounds when a test lead is connected to current input, but the selector switch is not set to a current range. Note for AC Measurements Model 2015 is a True-RMS meter. See Voltage Measurement section for further details. D • Gleich- und Wechselstrommessung (siehe Fig. 2) Anmerkung: Die Bereichsschaltung für Strommessungen ist manuell. “MANU” wird statt “AUTO” angezeigt. Œ Rotes Meßkabel für Messungen bis 400mA mit dem 400mA Eingang verbinden, oder für Messungen bis 20A mit dem 20A Eingang (20A nur für 30 Sekunden). Schwarzes Meßkabel mit COM verbinden.  Funktionsschalter auf gewünschten Bereich stellen. Auf A wenn Meßkabel mit 20A Eingang verbunden ist (mA Bereiche funktionieren nicht mit 20A Eingang); auf 4, 40 oder 400mA wenn Meßkabel mit 400mA verbunden ist. – 13 – 2015.qxd 7/21/00 5:40 PM Page 16 Fig. 2 Ž Strombereiche werden automatisch mit AC Funktion angewählt. “AC” wird angezeigt. Falls gewünscht, gelbe “Shift” Taste drücken um auf DC (Gleichstrom) überzuschalten.  Meßkreis abschalten, öffnen und Meßspitzen sicher in Serie mit dem Stromkreis verbinden.  Meßkreis einschalten. ‘ Stromwert ablesen. Mit 400mA Eingang von 400mA auf 40mA oder 4mA überschalten falls Auflösung zu niedrig ist. Wenn Sie Gleichstrom messen, fällt das Gerät bei Bereichswechsel zurück auf Wechselstrom. Gelbe “Shift” Taste drücken um wieder die Gleichstromfunktion zu wählen (“DC” Anzeige). – 14 – 2015.qxd 7/21/00 5:40 PM Page 17 Anmerkung 1: Beide Stromeingänge sind abgesichert. Sicherungen nur mit gleichwertigen ersetzen. Anmerkung 2 – Eingangswarnung: Wenn ein Meßkabel mit einem Stromeingang verbunden ist, der Funktionsschalter jedoch nicht auf einer Stromfunktion steht, wird ein akustisches Signal abgegeben. Eine solche Messung würde das Gerät zerstören. Anmerkung für Wechselstrommessung: Siehe Wechselspannungsmessung. E • Medidas de Corriente CC y CA (vea Fig. 2) Nota: la función de corriente no tiene selección automática de escala No aparece “AUTO” en el visualizador, sino “MANU”. Œ Conecte la punta de prueba roja a la entrada de 400 mA para medidas de corriente hasta 400 mA, o a la entrada de 20A para medidas hasta 20 A. Conecte la punta de prueba negra a la entrada COM.  Ponga el selector de función en la posición de corriente adecuada - en A cuando utilice la entrada de 20 A (las escalas de mA no funcionan con la entrada de 20 A), y en 4, 40 o 400 mA cuando utilice la entrada de 400 mA. Por defecto, al seleccionar una función de corriente, el multímetro se pone en modo AC (aparece “AC” en el visualizador). Pulse la tecla amarilla SHIFT para seleccionar el modo DC (aparece “DC” en el visualizador). Ž Abra el circuito sometido a prueba y conecte las puntas de prueba en serie con la carga.  Conecte la alimentación del circuito.  Lea el valor de la corriente en el visualizador. Si utiliza la entrada de 400 mA, cambie de 400 a 40 o 4 mA si es necesario. (Si está en modo DC, el medidor pasa a modo AC cada vez que gire el selector de función, por lo que deberá pulsar cada vez la tecla amarilla SHIFT para volver a DC). Nota 1: Cada entrada de corriente tiene un fusible de protección. En caso de que salte el fusible, cámbielo por otro igual. Nota 2 - Aviso de entrada incorrecta: El zumbador avisa si se ha conectado la punta de prueba a una entrada de corriente y el selector no está en una escala de corriente. Nota para medidas de CA El Modelo 2015 es un medidor de verdadero valor eficaz (TRMS): Vea “Medidas de tensión”. F • Mesure de Courant CC et CA (voir fig. 2) Note: Les gammes de courant sont commutées manuellement. Œ Connectez le cordon rouge à l’entrée 400mA pour mesures jusqu’à 400mA et à l’entrée 20A pour mesures jusqu’à 20A (20A pour 30 secondes). Connectez le cordon noir à l’entrée COM.  Placez le sélecteur – 15 – 2015.qxd 7/21/00 5:40 PM Page 18 sur le calibre souhaité. Sur “A”, quand l’entrée 20A est utilisée (les gammes mA ne fonctionnent pas avec l’entrée 20A), sur 4, 40, ou 400mA quand l’entrée 400mA est utilisée. Ž Une gamme de courant est automatiquement choisie en courant alternatif. Pressez la touche jaune SHIFT pour choisir courant continu.  Ouvrez le circuit à mesurer et connectez les pointes de touche solidement en série avec le circuit.  Mettez le circuit sous tension. ‘ Lisez la mesure. Avec l’entrée 400mA, descendez de gamme si la résolution est insuffisante. Si vous changez de gamme en mesure de courant continu, l’appareil se remet en courant alternatif. Pressez la touche jaune SHIFT pour vous remettre en courant continu (DC est affiché). Note 1: Les entrées de courant sont protégées par fusible. Ne remplacez les fusibles que par des fusibles équivalents. Note 2 – Avertissement d’entrée: Un signal sonore vous avertit quand un câble de mesure est connecté à une entrée de courant, alors que le sélecteur ne se trouve pas sur un calibre de courant. Une telle mesure détrouirerait l’appareil. Note pour la mesure de courant alternatif: Voir mesure de tension alternative. RESISTANCE MEASUREMENT (See Fig. 3) Œ Turn off any power to the resistance to be measured and discharge any capacitors. Any voltage present during a resistance measurement will cause inaccurate readings. Connect test leads as shown in figure 3. Ž Set Function/Range Switch to W position. Depress yellow button.  Connect test leads to circuit being measured.  Read resistance value on Digital Display. Open circuits will be displayed as an overload condition. D • Widerstandsmessung (siehe Fig. 3) Stellen Sie sicher daß keine Spannung am Widerstand oder Meßkreis anliegt. Eine Spannung würde die Messung verfälschen. Œ Meßkabel wie in Figur 3 verbinden.  Funktionsschalter auf W Position stellen. Gelbe Taste drücken. Ž Meßspitzen mit Schaltkreis verbinden.  Meßwert ablesen. Offene Schaltungen oder zu hohe Widerstände werden als Überlast angezeigt. E • Medidas de resistencia (vea Fig. 3) Asegúrese de que no hay tensión aplicada a la resistencia o al circuito. Œ Conecte las puntas de prueba como se muestra en la Figura 3.  Ponga el selector de función/escala en la posición W Ž Conecte las puntas de prueba al circuito sometido a medición.  Lea el valor de la resistencia en el visualizador. Un circuito abierto se indicará como sobrecarga. – 16 – 2015.qxd 7/21/00 5:40 PM Page 19 Press yellow button to select Resistance Gelben Knopf dr cken um Widerstand zu w hlen Pulse la tecla amarilla para seleccionar resistencia Pressez bouton jaune pour choisir RØsistance Fig. 3 F • Mesure de Résistance (voir fig. 3) Assurez-vous qu’aucune tension n’est appliquée à la résistance. Une tension fausserait le résultat. Œ Connectez les cordons comme indiqué en figure 3.  Placez le sélecteur sur W . Pressez la touche jaune. Ž Connectez les cordons au circuit à mesurer.  Lisez la valeur affichée. Un circuit ouvert ou une résistance trop élevée est affiché comme dépassement de calibre. CONTINUITY TEST (See Fig. 3) The Continuity test checks electrical continuity between two contact points. Œ Set the Function switch to W position.  Connect black test lead to COM input and red lead to V HzC input. Ž Connect probe tips to two circuit points.  The internal beeper emits a tone when resistance is less than approx. 40 W. – 17 – 2015.qxd 7/21/00 5:40 PM Page 20 D • Durchgangstest (siehe Fig. 3) Œ Funktionsschalter auf W -Bereich stellen.  Rotes Meßkabel mit V HzC Eingang und schwarzes mit COM Eingang verbinden. Ž Meßspizen mit Schaltkreis verbinden.  Akustisches Signal bei 100 W. E • Prueba de continuidad (vea Fig. 3) Œ Ponga el selector de función/escala en la posición W .  Conecte la punta de prueba negra a la entrada COM y la punta de prueba roja a la entrada V HzC. Ž Toque dos puntos del circuito con las puntas de prueba.  El zumbador suena cuando la resistencia es 100 W. F • Test de Continuité (voir Fig. 3) Œ Placez le sélecteur sur W .  Connectez le cordon rouge à l’entrée V HzC et le noir à l’entrée COM. Ž Connectez les pointes de touche au circuit.  Une résistance de 40 kW est indiquée par un signal sonore. DIODE TEST (See Fig. 4) The diode test measures the forward bias of a diode junction. Œ Connect the test leads as shown in figure 4.  Set the Function/range switch to . Depress yellow button. Ž Apply probe tip of red lead to the anode and of black lead to the cathode of the diode. The meter’s display indicates the forward voltage drop (approx. 0.6V for silicon diode or 0.4V for germanium diode). An open diode is indicated by a large voltage reading (>3.000 V).  Reverse test lead connections to the diode to perform a reverse bias test. A large value (>3.000 V) indicates a good diode. Notes: A large value (>3.000 V) for both reverse and forward bias tests indicates an open diode. A low voltage reading for both bias tests indicates a shorted diode. If the diode is shunted by a resistor of 1000 ohms or less, it must be removed from the circuit before taking the measurement. Bipolar transistor junctions may be tested in the same manner described above as emitter-base and base-collector junctions are diode junctions. D • Dioden- und Transistortest (siehe Fig. 4) Der Diodentest zeigt den Spannungsabfall über den Diodendurchgang Œ Meßkabel wie in Figur 4 verbinden.  Funktionsschalter auf stellen. Gelbe “Shift” Taste drücken. Ž Meßkabel mit Diode verbinden – rotes mit Anode; schwarzes mit Kathode. Spannungsabfall in Durchlaßrichtung ablesen (ung. 0.600V für eine Silikon-Diode und 0.400V für eine Ger- – 18 – 2015.qxd 7/21/00 5:41 PM Page 21 Fig. 4 maniumdiode. Eine offene Diode wird mit einem hohen Wert angezeigt (>3.000V).  Verbindung umdrehen um in Sperrrichtung zu messen. Eine hohe Ablesung (>3.000V) zeigt eine gute Diode an. Anmerkung: Eine hohe Ablesung (>3.000V) in beiden Richtungen zeigt eine offene Diode an; eine niedrige Ablesung eine kurzgeschlossene Diode. Transistorübergänge können wie Dioden getestet werden. E • Comprobación de diodos y transistores (vea Fig 4) En esta prueba se mide la polarización directa de la unión de un diodo. Œ Conecte las puntas de prueba como se muestra en la Figura 4.  Ponga el selector de función/escala en la posición Pulse la tecla amarilla. Ž Aplique la punta de prueba roja al ánodo y la negra al cátodo. El visualizador indica la caída de tensión directa (aprox. 0.6 V para diodos de Si, o 0.4 V para diodos de Ge). Una unión abierta se indica mediante una lectura muy elevada de tensión (>3.000 V).  Invierta la conexión de las puntas de prueba para comprobar la polarización inversa del diodo. Un valor elevado (>3.000 V) indica un diodo en buenas condiciones. – 19 – 2015.qxd 7/21/00 5:41 PM Page 22 Notas: Un valor elevado (>3.000 V) en ambos sentidos de polarización indica un diodo abierto. Un valor bajo en ambos sentidos indica un diodo cortocircuitado. Las uniones de un transistor bipolar son diodos, por lo que pueden comprobarse como tales. F • Test de Diodes et de Transistors (voir fig. 4) Le test de diodes affiche la chute de tension à travers le transistor. Œ Connectez les cordons comme indiqué en figure 4.  Placez le sélecteur sur . Pressez la touche jaune SHIFT. Ž Connectez les pointes de touche à la diode – le rouge à l’anode, le noir à la cathode.  Lisez la chute de tension en direction passante (environ 0.600V pour une diode au Si; 0.400V pour une diode au Ge. Une diode ouverte est affichée par une lecture élevée (>3.000V).  Inversez la connection pour mesurer en direction de bloquage. Une bonne diode est affichée par une lecture élevée (>3.000V). Notes: Une lecture élevée dans les deux directions indique une diode ouverte; une lecture basse indique une diode court-circuitée. Les jonctions de transistors peuvent être testées comme des diodes. CAPACITANCE MEASUREMENTS (See Fig. 5) Discharge all voltage from the capacitor before measurement. A safe way to discharge a capacitor is to connect a 100 kW resistor across the two capacitor leads. Œ Connect test leads as shown in figure 5.  Set the Function/Range switch to the range. Ž When measuring electrolytic capacitors, observe correct polarity (connect the COM test lead to the negative capacitor lead marked –).  Read the capacitance value on the display. D • Kapazitätsmessung (siehe Fig 5) Kondensator entladen (über 100 kW Widerstand). Œ Meßkabel wie in Figur 5 verbinden.  Funktionsschalter auf stellen. Ž Bei elektrolytischen Kondensatoren Polarität beachten.  Meßwert ablesen. E • Medidas de Capacidad (vea Fig. 5) Descargue la tensión presente en el condensador (a través de una resistencia de 100 K W). Œ Conecte las puntas de prueba como se muestra en la Figura 5.  Ponga el selector de función/escala en la posición . (Tenga en cuenta la polaridad di va a medir condensadores electrolíticos). Ž Lea el valor de la capacidad en el visualizador. – 20 – 2015.qxd 7/21/00 5:41 PM Page 23 Fig. 5 F • Mesure de Capacité (voir fig. 5) Déchargez le condensateur (à travers une résistance de 100k W ). Œ Connectez les cordons comme indiqué en figure 5.  Placez le sélecteur sur (respectez la polarité pour les condensateurs électrolytiques). Ž Lisez la valeur. FREQUENCY MEASUREMENT (See Fig. 6) Œ Connect the red test lead to the VW HzC Input and the black test lead to the COM Input.  Set the Function/Range switch to Hz position. Ž Connect test lead tips across the signal source.  Read the frequency value on the display. Note: Frequency counter operates in a 1000 count mode. Ignore numerals above decimals. D • Frequenzmessung (siehe Fig. 6) Œ Rotes Meßkabel mit V WHzC Eingang verbinden und schwarzes mit COM Eingang.  Funktionsschalter auf Hz stellen. Ž Meßspitzen mit Signalquelle verbinden.  Meßwert ablesen. Anmerkung: Frequenzmessung erfolgt in einem 1000-Punkte Anzeigenbereich. Die Ziffern über den Dezimalpunkten sind nicht zu beachten. – 21 – 2015.qxd 7/21/00 5:41 PM Page 24 E • Medidas de Frecuencia (vea Fig. 6) Œ Conecte la punta de prueba roja a la entrada V WHzC y la negra a la entrada COM.  Ponga el selector de función/escala en la posición Hz. Ž Conecte las puntas de prueba a la fuente de señal.  Lea el valor de la frecuencia en el visualizador. Nota: El medidor de frecuencia opera en un modo de 1000 cuentas. Ignore las cifras por encima de los decimales. F • Mesure de Fréquence (voir fig. 6) Œ Connectez le cordon rouge à l’entrée VW HzC et le noir à l’entrée COM.  Placez le sélecteur sur Hz. Ž Connectez les pointes de touche à la source de fréquence.  Lisez la valeur. Note: Pour la fréquence, l’affichage est limité à 1000 points. Les chiffres au-dessus des points décimaux sont à ignorer. Fig. 6 – 22 – 2015.qxd 7/21/00 5:41 PM Page 25 MIN/MAX MEASUREMENT Press MN/MX button to enter MIN Recording mode The meter switches to manual ranging with HOLD invoked. The last min. value is kept on the display until replaced by a lower reading. Press MN/MX button again to change to MAX recording mode. (MIN or MAX annunciator indicates what value is recorded). Overrange (flashing 4) is displayed when the range is exceeded. The bargraph continues to follow the measured signal. You can press the HOLD-button to stop recording, and press HOLD again to restart it. If recording is stopped, the minimum, maximum or present value as well as the bargraph are frozen. During MIN/ MAX recording mode Automatic Shutdown is disabled. With HOLD released, press RANGE button for two seconds to clear MX/MN and to return to autoranging mode. D • Min/Max Messung MN/MX Taste drücken um MIN Meßfunktion zu wählen. Das Multimeter schaltet nach manueller Bereichswahl mit aktivierter Anzeigesperre (HOLD). Die letzte niedrigste Messung wird auf der Anzeige festgehalten bis er durch einen niedrigeren Wert ersetzt wird. Erneut MN/MX Taste drücken um Maximale Werte festzuhalten. (MIN oder MAX Funktionsanzeige gibt an welcher Wert gemessen wird.) Bei Bereichsüberschreitung wird Überlast angezeigt. Während die Anzeige fesgehalten wird, folgt der Bargraf dem gemessenen Signal. Mit der HOLD Taste können Sie die Aufwertung stoppen und wieder starten (Min, Max, laufender Wert und Bargraf werden festgehalten). Während der MIN oder MAX Messung schaltet das Gerät nicht automatisch ab. Mit gelöster HOLD Taste die RANGE Taste zwei Sekunden lang drücken um die MX/MN Funktion zu verlassen und nach automatischer Bereichsschaltung zurückzukehren. E • Medidas Min/Max Pulse la tecla MN/MX para entrar en el modo de registro MIN. El medidor entra en selección manual de escala, con HOLD activada. El visualizador mantiene el último valor mínimo hasta que se registre un valor inferior. – 23 – 2015.qxd 7/21/00 5:41 PM Page 26 Pulse de nuevo MN/MX para cambiar al modo MAX. (El indicador MIN/MAX señala qué valor se está midiendo). Aparece la indicación de sobrecarga si se supera el fondo de escala. La barra analógica continúa siguiendo a la señal medida. Pulse de nuevo la tecla MN/MX para visualizar las medidas reales (el medidor permanece en modo de escala manual). Puede pulsar la tecla HOLD para detener el registro, y nuevamente HOLD para reanudarlo. Si se detiene el registro, tanto el valor mínimo, máximo o actual como la barra analógica quedan congelados. Durante el modo de registro MIN/MAX se anula la función de apagado automático. Tras liberar HOLD, pulse la tecla RANGE durante 2 segundos para cancelar MX/MN y volver al modo de escala automática. F • Mesure Min/Max Pressez la touche MN/MX pour choisir la fonction MIN. Le multimètre se met en sélection manuelle avec maintien de lecture (HOLD) activé. La dernière mesure la plus basse est gardée sur l’afficheur jusqu’à être remplacée par une nouvelle valeur inférieure. Pressez à nouveau la touche MN/MX pour afficher et maintenir le valeur la plus élevée. (MIN ou MAX est affiché selon la valeur mesurée.) Pendant que la valeur minimale ou maximale est affichée, le bargraphe suit le signal actuel. Avec la touche HOLD, vous pouvez arrêter la mesure et la résumer (Min, Max, la valeur courante et le bargraphe sont gelés). Pendant l’enregistrement MIN ou MAX, la coupure automatique est désactivée. Avec HOLD désactivé, pressez RANGE pendant deux secondes pour quitter la fonction MX/MN et retourner vers la sélection automatique. RELATIVE MODE With a measured value on the display, press REL to enter the Relative Mode. “REL D” appears on display, the displayed reading is memorized as reference value and the display resets to zero. For the next measurements, the display shows the difference between the new value and the stored reference value. Range Lock also works in Relative Mode. Press and hold down the REL-button for 2 seconds to exit the relative mode, or press RANGE for two seconds to clear REL and return to autoranging. – 24 – 2015.qxd 7/21/00 5:41 PM Page 27 D • Relativ-Messung Mit einem Meßwert auf der Anzeige, REL Taste drücken um RelativMessung zu wählen. “REL D” wird angezeigt. Der angezeigte Wert wird als Referenzwert gespeichert und die Anzeige wird auf Null gesetzt. Für folgende Messungen wird der Referenzwert vom gemessenen Wert abgezogen und die Differenz angezeigt. REL Taste oder RANGE Taste zwei Sekunden lang drücken um Relativ-Messung zu verlassen. E • Medidas Relativas Teniendo una medida en el visualizador, pulse REL para entrar en el modo de medidas relativas. Aparece “REL D“ en el visualizador, memorizándose la lectura presente como valor de referencia y reponiéndose el visualizador a cero. Para las medidas subsiguientes, el visualizador indica la diferencia entre el nuevo valor y el almacenado como referencia. En modo Relativo también funciona el bloqueo de escalas. Pulse y mantenga pulsada la tecla REL durante 2 segundos para salir del modo relativo, o bien RANGE durante 2 segundos para cancelar REL y volver al modo de escala automática. F • Mesure Relative Avec une mesure sur l’afficheur, pressez la touche REL. “REL D” est affiché. La valeur affichée est prise comme référence, et l’affichage est mis à zéro. Pour les prochaines mesures, cette valeur-référence est déduite du résultat, et seule la différence est affichée. Pressez la touche REL pendant deux secondes pour quitter la mesure relative. AUTOMATIC SHUTDOWN In order to save battery life, your multimeter turns off automatically after 30 minutes of inactivity. You can turn it back on by simply pressing the RESET button. You can disable the automatic shutdown by pressing and holding the RESET button while you initially switch on the instrument by turning the selector switch away from OFF to a desired measurement function. D • Automatische Abschaltung Ihr Multimeter schaltet nach 30 Minuten Inaktivität automatisch ab. Sie können es wieder einschalten durch die gelbe RESET Taste zu drücken. Sie können die automatische Abschaltung ausschalten durch beim Einschalten des Multimeters (Drehen des Funktionsschalters weg von OFF nach einer gewünschten Funtion) gleichzeitig die RESET Taste zu drücken. – 25 – 2015.qxd 7/21/00 5:41 PM Page 28 E • Apagado Automatico El multímetro se apaga automáticamente cuando transcurren 30 minutos sin actividad. Para volver a encenderlo basta con pulsar la tecla RESET. Es posible desactivar la función de apagado automático, pulsando y manteniendo pulsada RESET al mismo tiempo que enciende el instrumento, girando el selector a la posición deseada. F • Coupure Automatique Votre multimètre s’éteint automatiquement apres 30 minutes d’inactivité. Vous pouvez le ralummer en pressant la touche jaune RESET. Vous pouvez désactiver la coupure automatique en pressant la touche RESET au moment de mettre votre mutlimètre sous tension (tourner le sélecteur de OFF vers une fonction souhaitée). INCORRECT INPUT WARNING The Input Warning (beeper) is a feature to protect the meter and you from unintentional misuse. It warns you when a test lead is placed in a current input jack and the Selector switch is not set to current. (If a DMM is connected to a voltage test point with leads set for current, very high current could result). The warning reminds you that a test lead needs to be changed from a current jack to the voltage jack. All current ranges are fused with fast acting fuses. D • Eingangswarnung Ein akustisches Signal warnt Sie wenn ein Meßkabel mit einem Stromeingang verbunden ist, der Funktionsschalter jedoch nicht auf Strom steht. Eine solche Messung würde das Gerät zerstören. E • Aviso por Conexion Incorrecta de las Puntas de Prueba El zumbador suena si ha conectado una punta de prueba a una entrada de corriente y el selector no está en una escala de corriente. De seguir adelante con la medida podría dañar el multímetro. F • Avertissement d’Entrée Un signal sonore vous avertit quand un cordon de mesure est connecté à une entrée de courant alors que le sélecteur n’est pas placé sur une gamme de courant. Une telle mesure endommagerait l’appareil. – 26 – 2015.qxd 7/21/00 5:41 PM Page 29 SAFETY TEST LEADS The test leads included with your meter have shrouded banana plugs to eliminate the possibility of shock if the plugs accidentally pull out of the meter while making a measurement. The test leads also have insulated probe tips to avoid causing shorts when making measurements in dense component areas. Replacement part number for safety test leads is TL245. D • Sicherheitsmeßkabel Die Meßkabel haben versenkte Bananenstecker um elektrischen Schock zu vermeiden. Die Meßspitzen sind zum Teil isoliert, um Kurzschlüsse in dichten Schaltungen zu vermeiden. Diese Isolation kann entfernt werden. E • Puntas de Prueba de Seguridad Las puntas de prueba suministradas con el multímetro incluyen unos conectores de banana protegidos para eliminar la posibilidad de descargas eléctricas. Las puntas están parcialmente aisladas para evitar cortocircuitos en áreas con alta densidad de componentes. El usuario puede quitar dicho aislante si lo desea. F • Cordons de Sécurité Les fiches banane des cordons sont munis de protecteurs fixes afin de supprimer les risques de chocs électriques. Les pointes métalliques sont partiellement isolées pour éviter des court-circuits dans des circuits denses. Cette isolation peut être enlevée. – 27 – 2015.qxd 7/21/00 5:41 PM Page 30 SPECIFICATIONS General Specifications Display: 3-3/4 digit LCD, 4000 counts, 9999 counts in Frequency mode, 40 segment bargraph with function and unit annunciators. Polarity Indication: Automatic, positive implied, negative (–) indicated Zero Adjustment: Automatic Overrange Indication: Most significant digit (MSD) blinks. Other digits shown as zero. Low Battery Indication: . Change battery immediately. Display Update Rate: 2/sec, nominal; 1/sec in capacitance and frequency modes; 20/sec for bargraph. Operating Temp.: 0 to 50°C, 0 to 70% Relative Humidity Storage Temp: -20°C to 60°C, 0 to 80% RH, battery removed Temperature coefficient: 0.1 x (spec. accuracy)/°C (<18°C or >28°C° Power: Standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22 Auto Power-Off: Meter turns off after 30 minutes of inactivity. Battery Life (Typical): 500 hours alkaline, 250 hours carbon-zinc Dimensions (H x W x D): 7.5” x 3.5” x 1.3”, (175 x 90 x 34 mm) Weight (incl. battery): 568 g (20 oz) Accessories: Test leads, spare fuse, battery, and Operator’s Manual Safety: Meets EN61010-1 Cat III 1000V; EMC: Meets EN50081-1, EN50082-2. EMC: This product complies with requirements of the following European Community Directives: 89/336/EEC (Electromagnetic Compatibility) and 73/23/EEC (Low Voltage) as amended by 93/68/EEC (CE Marking). However, electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement circuit. Measuring instruments will also respond to unwanted signals that may be present within the measurement circuit. Users should exercise care and take appropriate precautions to avoid misleading results when making measurements in the presence of electronic interference. Electrical Specifications Accuracies at 23°C ± 5°C, <75% RH. DC Volts Ranges: 400mV, 4, 40, 400, 1000V Resolution, 100µV in 400mV range: Accuracy, all Rgs: ±(0.25%rdg +2dgt) Input Impedance, 400mV rng. >1000 MW ; 4V rng: 11 MW ; 40V1000V rngs: 10 M W OL Protection: 1100VDC or Peak AC. 6kV transient protection. AC Volts (50-1000Hz). Model 2015: True RMS meter. Accuracies apply to all waveforms. Ranges: 400mV, 4, 40, 400, 750V Resolution: 100µV in 400mV range Crest Factor 3:1 Accuracy ±(%rdg +nr digits): Range 50-500Hz 500-1000Hz 400mV* ±(1.0% +5) unspecified 40V-400V ±(0.8% +3) ±(1.0% +5) All other ±(0.8% +3) ±(1.2% +5) Input Impedance, 400mV rg. >1000 MW ; 4V rg: 11 MW ; 40V-750V rgs: 10 M W OL Protection: 1100VDC or Peak AC. 6kV transient protection. – 28 – 2015.qxd 7/21/00 5:41 PM * Frequency response 50-100Hz DC Current Ranges: 4, 40, 400mA, 20A Resolution, 4mA range: 1µA Accuracy, 4mA to 400mA ranges: ±(0.8%rdg + 2dgt) 20A range: ±(1.2%rdg ±4dgt) Voltage burden: 500mV. 400mA rg.: 900mV OL Protection: mA input: 0.5A/250V fast blow ceramic fuse (FP250, 6.4x25.4mm); 10kA interrupt rating 20A input: 10A continuous, 20A for 30 seconds. Fuse: Littel Fuse KLK 20A or Bussman KTK 20A 10.3x38.1mm fast acting fuse. 100kA interrupt rating. AC Current (50-1000Hz) Model 2015: True RMS meter. Accuracies apply to all waveforms. Ranges: 4, 40, 400mA, 20A Resolution, 4mA range: 1µA Crest Factor 3:1 Accuracy, 4mA to 400mA ranges: ±(1.2%rdg + 4dgt) 20A range: ±(1.5%rdg ±5dgt) Voltage burden: see DC current OL Protection: see DC current. Range: 4V Resolution: 1mV Test current: 0.6mA Display: Forward Junction Voltage Accuracy: ±(1.0%rdg +1dgt) Open Circuit Voltage: 3.2Vdc Overload prot.: 500VDC or RMS AC Continuity Indicator Range: 400 Ohms Audible threshold: < 40 W Response time: approx. 100 ms Open circuit voltage: 400 mV DC OL Protection: 500 VDC or RMS AC Capacitance Ranges: 4, 40, 400nF, 4, 40µF Accuracy, 4 nF range: ±(1.0%rdg + 40dgt); 40 nF-4µF: ±(1.0%rdg + 4dgt); 40 µF (<20µF): ±(1.0%rdg + 4dgt); 40 µF (>20 µF): ±(5.0%rdg + 8dgt) Test frequency, all ranges: 1Hz Note: use relative mode for 4nF and 40nF ranges. Resolution, 4nF range: 0.001nF OL Protection: 500 VDC or RMS AC Frequency Counter Resistance Range/Accuracy: 400 W /±(0.8%rdg + 4dgt) 4 k- 400 kW /±(0.7%rdg + 2dgt); 4 MW /±(1.0%rdg + 4dgt); 40 M W/±(1.5%rdg + 4dgt) Resolution, 400 W range: 100 m W Test voltage, all ranges: 0.4VDC OL protection, all Ranges: 500VDC or AC RMS Diode Test Page 31 Ranges: 100 Hz, 1, 10, 100, 400 kHz Accuracy, 100 Hz rg: ±(0.1%rdg +10dgt); 1 & 10 kHz: ±(0.1%rdg +4dgt); 100 kHz: ±(0.1%rdg +8dgt); 400 kHz: ±(0.1%rdg +20dgt) Resolution, 100Hz range: 0.01Hz Trigger Level - 100 Hz-100 kHz: 40 mV; 100kHz-400kHz: 400mV OL Protection: 500VDC or RMS AC Note: For frequencies below 1Hz, display indicates 00.00Hz. Optional Accessories RF241 DL243C DL248C TL253A TL245 CT231A CT232A CT234A – 29 – 650MHz RF Probe Basic Test Lead Set Deluxe Test Lead Set Temperature Converter (900°C/1652°F) Replacement Safety Test Leads 150A AC Current Clamp 1000A AC Current Clamp 400A AC Current Clamp 2015.qxd CT235 CT236A CT237 CT238 VC221A D 7/21/00 5:41 PM 1000A AC/DC Clamp 500A AC Clamp (mV output) 200A AC/DC Current Clamp 20A AC/DC Current Clamp Padded Vinyl Case. Fits meter with holster Page 32 VC231 Padded Vinyl Case. Fits meter without holster. DC205C Deluxe Hard-Shell Carry Case DC207C Large Deluxe Hard-Shell Carry Case with extra space for accessories SPEZIFIKATIONEN Allgemeine Spezifikationen Anzeige: 3-3/4 Digit LCD, 4000 Punkte, 9999 Punkte bei Frequenzmessung, 40-Segment Bargraf mit Funktions- und Einheitsanzeigen. Polaritätsanzeige: Automatisch, positiv unterstellt, negativ (–) angezeigt. Nullabgleich: Automatisch Überlastanzeige: Höchste Ziffer (4) blinkt. Andere als Null angezeigt. Entladene Batterieanzeige: . Batterie sofort wechseln. Anzeigenauffrischung: 2/Sek, nominal; 1/Sek für Kapazität und Frequenz; 20/Sek für Bargraf. Betriebstemp.: 0 bis 50°C, 0 bis 70% Relative Feuchtigkeit Lagertemp: -20°C bis 60°C, 0 bis 80% R.F., Batterie entfernt. Temperaturkoeffizient: 0.1 x (angegebene Genauigkeit)/°C (<18°C oder >28°C° Stromversorgung: Standard 9-Volt Transistorbatterie, NEDA 1604, JIS 006P, IEC 6F22 Automatische Abschaltung: Nach 30 Minuten Inaktivität. Autonomie (typisch): 500 Stunden Alkaline, 250 Stunden Zinn-Kohle Abmessungen (H x B x T): 175 x 90 x 34 mm Gewicht mit Batterie: 568 g Zubehör: Meßkabel, Ersatzsicherung, Batterie, Anleitung Sicherheit: Gemäß EN61010-1 Cat.III 1000V; EMC: Gemäß EN50081-1, EN50082-2 EMC Dieses Produkt beantwortet an die Bestimmungen der folgenden EWG Richtlinien: 89/336/EEC (Elektromagnetische Kompatibilität) und 73/23/EEC (Niedrige Spannung) geändert durch 93/68/EEC (CE Marking). Elektrisches Rauschen und starke magnetische Felder in der direkten Umgebung des Meßgerätes können jedoch den Meßkreis beeinflussen. Das Gerät kann auch durch Störsignale im gemessenen Schaltkreis beeinflußt werden. Der Anwender muß Vorsichtsmaßnahmen treffen um irreführende Meßergebnisse bei Messungen in der Umgebung von starken elektromagnetischen Feldern zu vermeiden. Elektrische Spezifikationen Genauigkeit bei 23°C ± 5°C, <75% R.F. Gleichspannung Bereiche: 400mV, 4, 40, 400, 1000V Auflösung, 100µV im 400mV Ber. Genauigkeit, alle Bereiche: ±(0.25%vMW +2Dgt) Eingangsimpedanz, 400mV Bereich >1000MW, 4V Ber: 11M W; 40V1000V Ber.: 10MW Überlastschutz: 1100VDC oder AC Spitze. 6kV Transienten. Wechselspannung (50-1000Hz). Modell 2015: Echt-Effektivwert Messung. Bereiche: 400mV, 4, 40, 400, 750V – 30 – 2015.qxd 7/21/00 5:41 PM Auflösung: 100µV im 400mV Ber. Crest Faktor: 3:1 Genauigk. ±(%Bereich +Anz. Digits): Bereich 50-500Hz 500-1000Hz 400mV* ±(1.0% +5) nicht spezifiz. 40-400V ±(0.8% +3) ±(1.0% +5) Andere ±(0.8% +3) ±(1.2% +5) Eingangsimpedanz, 400 mV Ber. >1000 M W; 4V Ber.: 11 MW ; 40V750V Ber.: 10 MW Überlastschutz: 1100VDC oder AC Spitze. 6kV Transienten. * Frequenzgang: 50-100Hz Gleichstrom Bereiche: 4, 40, 400mA, 20A Auflösung, 4mA Bereich: 1µA Genauigkeit, 4mA bis 400mA Bereiche: ±(0.8%vMW + 2Dgt) 20A Bereiche: ±(1.2%vMW ±4Dgt) Spannungsabfall: 500mV, ausg. 400mA Bereich.: 900mV Überlastschutz: mA Eingang: 0.5A/250V schnelle Sicherung (FP250 - 6.4 x 25.4mm); 10kA Trennvermögen 20A Eingang: 10A kontinuierlich, 20A für 30 Sekunden. Littel Fuse KLK 20A oder Bussman KTK 20A 10.3x38.1mm flinke Sicherung. 100kA Trennvermögen. Page 33 4 M W/±(1.0%vMW + 4Dgt); 40M W/±(1.5%vMW + 4Dgt) Auflösung, 400 W Bereich: 100mW Test Spannung, alle Bereiche: 0.4V Überlastschutz, alle Bereiche: 500VDC oder AC effektiv Diodentest Bereich: 4V Aulösung: 1mV Teststrom: 0.6mA Anzeige: Spannungsabfall in Durchlaßrichtung Genauigkeit: ±(1.0%vMW +1Dgt) Leerlaufspannung: 3.2VDC Überlastschutz: 500VDC oder AC effektiv Durchgangstest Bereich: 400 Ohms Akustisches Signal bei: < 40 W Ansprechzeit: ung. 100ms Leerlaufspannung: 400mV DC Überlastschutz: 500VDC oder RMS effektiv Kapazität Wechselstrom (50-1000Hz) Modell 2015: Echt-Effektivwert Messung. Bereiche: 4, 40, 400mA, 20A Auflösung, 4mA Bereich: 1µA Crest Faktor: 3:1 Genauigkeit, 4mA bis 400mA Bereiche: ±(1.2%vMW + 4Dgt) 20A Bereich: ±(1.5%vMW ±5Dgt) Spannungsabfall: siehe Gleichstrom Überlastschutz: siehe Gleichstrom Bereiche: 4-400nF, 4-40µF Genauigkeit, 4nF Bereich: ±(1.0%vMW + 40Dgt); 40nF-4µF: ±(1.0%vMW + 4Dgt); 40µF (<20µF): ±(1.0%vMW + 4Dgt); 40µF (>20µF): ±(5.0%vMW + 8Dgt) Test Frequenz, alle Bereiche: 1Hz Anmerkung: Relativmessung für 4nF und 40nF Bereiche verwenden Auflösung, 4nF Bereich: 0.001nF Überlastschutz: 500VDC oder AC effektiv Frequenzzähler Widerstand Bereich/Genauigkeit: 400 W /±(0.8%vMW + 4Dgt) 4k-400 kW/±(0.7%vMW + 2Dgt); Bereiche: 100Hz, 1, 10, 100, 400kHz Genauigkeit, 100Hz Bereich: ±(0.1%vMW +10Dgt); 1 & 10kHz: ±(0.1%vMW +4Dgt); 100kHz: ±(0.1%vMW +8Dgt); 400kHz: ±(0.1%vMW +20Dgt) – 31 – 2015.qxd 7/21/00 5:41 PM Auflösung, 100Hz Bereich: 0.01Hz Überlastschutz: 500VDC oder AC effektiv Anmerkungen: Für Frequenzen unter 1Hz, wird 00.00Hz angezeigt. Optionszubehör RF241 DL243C DL248C TL253A 650MHz Hochfrequenzsonde Standard Testkabelsatz Deluxe Testkabelsatz Temperaturumsetzer (900°C/1652°F) TL245 Ersatz-Sicherheitsmeßkabel CT231A 150A Wechselstromzange CT232A 1000A Wechselstromzange E Page 34 CT234A 400A Wechselstromzange CT235 1000A AC/DC Stromzange CT236 500A Wechselstromzange (mV Ausgang) CT237 200A AC/DC Stromzange CT238 20A AC/DC Stromzange VC221A Gefütterte Vyniltasche. Für Meter und Holster VC231 Gefütterte Vyniltasche. Für Meter ohne Holster DC205C Deluxe-Hartmantel-Tragetasche DC207C Große Deluxe-HartmantelTragetasche mit mehr Platz für Zubehör ESPECIFICACIONES Especificaciones Generales Visualizador: LCD de 3-3/4 dígitos, 4000 cuentas, 9999 cuentas en medidas de frecuencia, barra analógica de 40 segmentos, indicadores de función y unidades Indicación de polaridad: Automática, positiva implícita, negativa indicada Ajuste de cero: Automático Indicación de sobrecarga: Dígito más significativo intermitente, otros dígitos fijos a cero Indicación de “pila baja”: . Esta indicación señala el momento de cambiar la pila. Frecuencia de refresco de la lectura: 2 veces/segundo nominal; 1 vez/s en medidas de capacidad y frecuencia; 20 veces/s la barra analógica. Temp. de funcionamiento: 0 a 50 °C, 0 a 70% H.R. Temp. de almacenamiento: -20 a 60 °C, 0 a 80% H.R., sin pila. Coef. de temperatura: 0.1 x (precisión espec.)/°C (<18 °C o >28 °C) Alimentación: Pila normal de 9 V, NEDA 1604, JIS 006P, IEC 6F22 Apagado automático: El medidor se apaga transcurridos 30 minutos sin actividad Duración de la pila (típica): alcalina 500 hrs, carbono-zinc 250 hrs Dimensiones: 175 x 90 x 34 mm Peso (pila incluida): 568 g Accesorios: Puntas de prueba, fusible de repuesto, pila y Manual de Instrucciones Seguridad: Según normas EN61010-1 Cat.III 1000V; EMC: Según EN50081-1, EN50082-2 EMC: Este producto cumple los requisitos de las siguientes Directivas de la Comunidad Europea: 86/336/EEC (Compatibilidad Electromagnética) y 73/23/EEC (Baja Tensión), con enmiendas según 93/68/EEC (Marcado CE). No obstante, la presencia de ruido eléctrico o campos electromagnéticos intensos en las proximidades del equipo pueden introducir perturbaciones en los circuitos de medida. Los instrumentos de medida también responden a las señales no deseadas que puedan estar presentes en los circuitos de medida. El usuario deberá tomar las precauciones necesarias para evitar obtener resultados incorrectos cuando realiza medidas en presencia de interferencias electromagnéticas. – 32 – 2015.qxd 7/21/00 5:41 PM Especificaciones eléctricas Valores precisión a 23 ºC ±5 ºC, H.R. <75% Voltios CC Page 35 man KTK 20A 10.3x 38.1mm. Nivel de interrupción 100 KA. Corriente CA (50 - 1000 Hz) Modelo 2015: Medidor de verdadero valor eficaz - precisión aplicable a todas las formas de onda Escalas: 400 mV; 4, 40, 400, 1000 V Resolución: 100 mV en esc. de 400 mV Precisión (todas las escalas): ±(0.25% lect. +2 dígitos) Imped. de entrada: Esc. 400 mV: >1000MW ; escala 4 V: 11MW; escalas 40-1000 V: 10MW Protecc. sobrec.: 1100 V CC o pico CA. Protección transitorios 6 KV Voltios CA (50 - 1000 Hz) Modelo 2015: Medidor de verdadero valor eficaz - precisión aplicable a todas las formas de onda Escalas: 400 mV; 4, 40, 400, 750 V Resolución: 100 mV en esc. de 400 mV Factor Cresta: 3:1 Precisión: ±(% escala +nº de dígitos): Escala 50-500 Hz 500-1000 Hz 400 mV* ±(1.0% +5) sin especificar 40-400 V ±(0.8% +3) ±(1.0% +5) Otras ±(0.8% +3) ±(1.2% +5) Impedancia de entrada: escala 400mV: >1000 MW ; escala 4 V: 11 MW; esc. 40-750 V: 10 MW Protecc. sobrec.: 1100 V CC o pico AC. Protección transitorios 6 KV *Respuesta en frecuencia 50-100 Hz Corriente CC Escalas: 4, 40, 400 mA; 20 A Resolución: 1 mA, escala de 4 mA Precisión: escalas 4 mA- 00 mA: ±(0.8% lect +2 dgt); escala 20 A: ±(1.2% lect ±4 dgt) Lastre de tensión: 500 mV. Escala 400 mA: 900 mV Protección sobrecorriente: Entrada mA: fusible rápido de 0.5A/250V (FP250 - 6.4 x 25.4mm); nivel de interrupción 10 KA Entrada A: 10 A continuo, 20 A durante 30 s. Fusible cerámico rápido Littel Fuse KLK 20A o Buss- Escalas: 4, 40, 400 mA; 20 A Resolución: 1 mA en la esc. de 4 mA Factor Cresta: 3:1 Precisión: escalas 4 mA a 400 mA: ±(1.2% lect. +4 dgt); escala 20 A: ±(1.5% lect ±5 dgt) Lastre de tensión: vea Corriente CC Protecc. sobrec.: vea Corriente CC Resistencia Escalas/Precisión: 400 W /±(0.8% lect +4 dgt) 4 k-400 K W/±(0.7% lect +2 dgt) 4 M W/±(1.0% lect +4 dgt) 40 M W/±(1.5% lect +4 dgt) Resolución, escala 400 W : 100 m W Tensión de prueba: todas esc.: 0.4V Protección sobrecarga, todas las escalas: 500 V CC o CA ef. Prueba de diodos Escala: 4 V Resolución: 1 mV Precisión: ±(1.0% lect +1 dgt) Corriente de prueba: 0.6 mA Indicación: Tensión directa de la unión Tensión circuito abierto: 3.2 V CC Protecc. sobrec.: 500 V CC o CA ef. Indicador de continuidad Escala: 400 W Umbral audible: <40 W Tiempo de respuesta: aprox. 100 ms Tensión, circuito abierto: 400 mV CC Protecc. sobrec.: 500 V CC o CA ef. Capacidad Escalas: 4-400 nF, 4-40 µF Precisión, escala 4 nF: ±(1.0% lect +40 dgt); 40 nF-4 µF: ±(1.0% lect +4 dgt); 40 µF (<20 µF): ±(1.0% lect +4 dgt); 40 µF (>20 µF): ±(5% – 33 – 2015.qxd 7/21/00 5:41 PM lect +8 dgt) Frecuencia de prueba, todas esc. 1Hz Nota: utilice el modo relativo para las escalas de 4 nF y 40 nF Resolución, escala 4 nF: 0.001 nF Protecc. sobrec.: 500 V CC o CA ef. Contador de frecuencia Escalas: 100 Hz; 1, 10, 100, 400 KHz Precisión, escala 100 Hz: ±(0.1% lect +10 dgt) 1 y 10 KHz: ±(0.1% lect +4 dgt) 100 KHz: ±(0.1% lect +8 dgt) 400 KHz: ±(0.1% lect +20 dgt) Resolución, escala 100 Hz: 0.01 Hz Nivel de disparo, 100 Hz-100 KHz: 40 mV; 100 KHz-400 Kz: 400 mV Protecc. sobrec.: 500 V CC o CA ef. Nota: Para frecuencias por debajo de 1 Hz, el visualizador indica 00.00 Hz. Page 36 DL248C Juego de puntas de prueba (especial) TL253A Convertidor de temperatura (900°C/1652°F) TL245 Puntas de prueba de seguridad de repuesto CT231A Pinza de corriente 150 A CA CT232A Pinza de corriente 1000 A CA CT234A Pinza de corriente 400 A CA CT235 Pinza 1000 A CA/CC CT236A Pinza 500 A CA (salida mV) CT237 Pinza de corriente 200 A CA/CC CT238 Pinza de corriente 20 A CA/CC VC221A Estuche de vinilo acolchado. Cabe el medidor con funda protectora VC231 Estuche de vinilo acolchado. Cabe el medidor sin funda protectora DC205C Estuche de lujo con carcasa dura DC207C Estuche de lujo grande con carcasa dura y con espacio extra para accesorios Accesorios opcionales RF241 Sonda de RF 650 MHz DL243C Juego de puntas de prueba (básico) F SPECIFICATIONS Spécifications Générales Afficheur: LCD 3-3/4 digits, 4000 points (9999 points en mode Fréquence), avec bargraphe à 40 segments, indicateurs de fonctions et d’unités. Indication de Polarité: Automatique, positive impliquée, négative (–) indiquée. Adjustement du zéro: Automatique Indication de dépassement de gamme: Chiffre plus significatif “4” clignotte. Autres chiffres indiqués comme zéro. Indication de pile déchargée: . Changez la pile immédiatement Raffraichissement: 2/sec, nominal; 1/sec en modes capacité et fréquence; 20/sec pour le bargraphe. Temp. de Fonctionnement: 0 à 50°C, 0 à 70% Humidité Relative Temp. de Stockage: -20°C à 60°C, 0 à 80% H.R., pile enlevée Coefficient de Température: 0.1 x (précison spécifiée)/°C (<18°C ou >28°C° Alimentation: Pile 9-Volt Standard, NEDA 1604, JIS 006P, IEC 6F22 Coupure Automatique: après 30 minutes d’inactivité. Autonomie (typique): 500 heures alcaline, 250 heurs zinc-charbon Dimensions (H x L x P): 175 x 90 x 34 mm Poids (avec pile): 568 g Accessoires: Cordons de test, fusible de réserve, pile, manuel d’utilisation – 34 – 2015.qxd 7/21/00 5:41 PM Sécurité: Conforme à EN61010-1 Cat.III 1000V; EMC: selon EN50081-1, EN50082-2 EMC: Ce produit est conforme aux exigences des directives suivantes de la Communauté Européenne: 89/336/EEC (Compatibilité Electromagnétique) et 73/23/EEC (Basse Tension), modifiée par 93/68/EEC (CE Marking). Cependant, du bruit électrique ou des champs électromagnétiques intenses dans la proximité de l’instrument peuvent influencer le circuit de mesure. L’instrument peut également être perturbé par des signaux parasytes dans le circuit mesuré. L’utilisateur doit être vigilant et prendre des précautions appropriées pour éviter des résultats erronés quand les mesures sont prises en présence d’interférences électromagnétiques. Spécifications Electriques Précisions à 23°C ± 5°C, <75% H.R. Tension Continue Gammes: 400mV, 4, 40, 400, 1000V Résolution, 100µV en gamme 400mV: Précision, toutes gammes: ±(0.25%lect +2dgt) Impédance d’entrée, gm 400mV: >1000MW ; gm. 4V: 11M W; gms. 40V-1000V: 10MW Protection de surcharge: 1100Vcc ou crête ca. Transitoires de 6kV. Page 37 Factor crête: 3:1 Précision ±(%gamme +nbr digits): Gamme 50-500Hz 500-1000Hz 400mV* ±(1.0% +5) non spécifié 40V-400V ±(0.8% +3) ±(1.0% +5) Autres ±(0.8% +3) ±(1.2% +5) Impédance d’entrée, gm. 400mV >1000WM ; gm 4V: 11MW ; gms 40V-750V: 10MW Protection de surcharge: 1100Vcc ou crête ca. Transitoites de 6kV. * Réponse Fréquence 50-100Hz Courant Continu Gammes: 4, 40, 400mA, 20A Résolution, gamme 4mA: 1µA Précision, gammes 4mA à 400mA: ±(0.8%lect + 2dgt) gamme 20A: ±(1.2%lect ±4dgt) Chute de tension: 500mV. Gamme 400mA: 900mV Protection de surcharge: Entrée mA: fusible rapide 0.5A/250V (FP250 6.4 x 25.4 mm); pouvoir de coupure 10kA Entrée 20A: 10A en continu, 20A pour 30 seconds. Fusible rapide Littel Fuse KLK 20A ou Bussman KTK 20A 10.3x38.1mm. 100kA interrupt rating.Pouvoir de coupure: 100kA. Courant Alternatif (50-1000Hz) Modèle 2015: Mesure efficace vraie. Les précisions s’appliquent à toutes les formes d’onde. Gammes: 4, 40, 400mA, 20A Résolution, gamme 4mA: 1µA Factor crête: 3:1 Précision, gammes 4mA à 400mA: ±(1.2%lect + 4dgt) Tension Alternative (50-1000Hz). gamme 20A: ±(1.5%lect ±5dgt) Modèle 2015: Mesure efficace vraie. Les Chute de tension: voir courant cc. précisions s’appliquent à toutes les Protection de surcharge: voir courant formes d’onde. cc. Gammes: 400mV, 4, 40, 400, 750V Résolution: 100µV en gamme 400mV. – 35 – 2015.qxd 7/21/00 5:41 PM Page 38 1 & 10kHz: ±(0.1%lect +4dgt); 100kHz: ±(0.1%lect +8dgt); 400kHz: ±(0.1%lect +20dgt) Résolution, gamme 100Hz: 0.01Hz Niveau de déclenchement - 100Hz100kHz: 40mV; 100kHz-400kHz: 400mV Protection de surcharge: 500Vcc ou ca effectif Résistance Gamme/Précision: 400 W/±(0.8%lect. + 4dgt) 4k-400kW /±(0.7%lect + 2dgt); 4MW /±(1.0%lect + 4dgt); 40MW /±(1.5%lect + 4dgt) Résolution, gamme 400W : 100mW Tension de test, tts gammes: 0.4V Protection de surcharge, toutes gammes: 500VDC ou AC RMS Note: Pour fréquences <1Hz, affichage de 00.00Hz. Test de Diodes Gamme: 4V Résolution: 1mV Courant de test: 0.6mA Affichage: Chute de tension en direction passante Précision: ±(1.0%lect +1dgt) Tension en circuit ouvert: 3.2Vdc Protection de surcharge: 500Vcc ou ca effectif. Indication de Continuité Gamme: 400 Ohms Seuil pour indication sonore: <40 W Temps de réponse: approx. 100ms Tension en circuit ouvert: 400mVcc Protection de surcharge: 500Vcc ou ca effectif. Capacité Accessoires en Option RF241 DL243C DL248C TL253A TL245 CT231A CT232A CT234A CT235 CT236A CT237 CT238 VC221A VC231 Gammes: 4, 40, 400nF, 4, 40µF Précision, gamme 4nF: ±(1.0%lect + 40dgt); 40nF-4µF: ±(1.0%lect + 4dgt); 40µF (<20µF): ±(1.0%lect + 4dgt); 40µF (>20µF): ±(5.0%lect + 8dgt) Fréquence test, tts gammes: 1Hz DC205C DC207C Note: utilisez le mode relatif pour gammes 4nF et 40nF Résolution, gamme 4nF: 0.001nF Protection de surcharge: 500Vcc ou ca effectif Compteur de Fréquence Gammes: 100Hz, 1, 10, 100, 400kHz Précision, gm. 100Hz: ±(0.1%lect +10dgt); – 36 – Sonde RF 650MHz Set de cordons standards Set de cordons deluxe Convertisseur de Températ. (900°C/1652°F) Cordons de remplacement Pince de courant 150A ca Pince de courant 1000A ca Pince de courant 400A ca Pince de courant 1000A ca/cc Pince de courant 500A ca (sortie mV) Pince de courant 200A ca/cc Pince de courant 20A ca/cc Sacoche vynil rembourrée. Pour instrument avec holster. Sacoche vynil rembourrée. Pour instrument sansholster. Etui rigide de luxe Large étui rigide de luxe avec espace de rangement pour les accessoires 2015.qxd 7/21/00 5:41 PM Page 39 TROUBLESHOOTING and MAINTENANCE If there appears to be a malfunction during the operation of the meter, the following steps should be performed in order to isolate the cause of the problem: Œ Check the battery.  Review the operating instructions for possible mistakes in operating procedure. Ž Inspect and test the Test Probes for a broken or intermittent connection.  Inspect and test the fuse. See Fuse Replacement. Except for the replacement of the battery or fuse, repair of the multimeter should be performed only by a Factory Authorized Service Center or by other qualified instrument service personnel. The front panel and case can be cleaned with a mild solution of detergent and water. Apply sparingly with a soft cloth and allow to dry completely before using. Do not use aromatic hydrocarbons or chlorinated solvents for cleaning. D • Fehlersuche und Unterhalt Prüfen Sie zuerst folgende Fehlerquellen: Meßkabel (Brüche), Anschluß, Zustand von Batterie und Sicherung, richtiger Meßvorgang, Eingangs- und Bereichsgrenzen, usw. Mit Ausnahme des Batterie- und Sicherungswechsels soll jede Reparatur nur durch eine durch Wavetek anerkannte Servicestelle durchgeführt werden. Das Gerät kann mit einer milden Seifenlösung gereinigt werden. Sparsam auftragen und vor Gebrauch gut trocknen lassen. Keine Lösungsmittel zum Reinigen verwenden. E • Mantenimiento y reparación Para identificar la causa del problema: Compruebe la pila; revise las instrucciones de uso; inspeccione las puntas de prueba por si hay una conexión rota o intermitente; inspeccione la pila y el fusible. Excepto la sustitución de la pila o el fusible, cualquier trabajo de reparación del multímetro debe hacerse exclusivamente por personal técnico cualificado para este tipo de reparaciones. Para limpiar la carcasa puede utilizarse una solución suave de agua y detergente. F • Maintenance et Réparation Avant d’expédier votre multimètre pour réparation, vérifiez les cordons de mesure (rupture), pile et fusible, connections, procédure de mesure, limites d’entrée et de calibres, etc. – 37 – 2015.qxd 7/21/00 5:41 PM Page 40 Excepté pour le remplacement de la pile et du fusible, toute réparation doit être effectuée uniquement par un Centre de Services agrée par Wavetek. Vous pouvez nettoyer le boîtier avec un détergent doux. Appliquez parcimonieusement et laissez sécher complètement avant utilisation. Ne pas utiliser de solvents. BATTERY / FUSE REPLACEMENT (Fig. 7) Warning: To prevent electrical shock hazard, turn off the multimeter and any device or circuit under test and disconnect the test leads before removing the rear cover. Fig. 7 Input Eingang Entrada Entrée Fuse Sicherung Fusible Fusible Type Typ Tipo Type Dimensions Abmessungen Dimensiones Dimensions mA 0.5A/250V fast blow ceramic fuse1) FP250 6.4x25.4mm 20A 20A/600V fast blow ceramic fuse1) Littel Fuse KLK 20A 10.3x38.1mm Bussman KTK 1) Interrupt rating Trennvermögen Niv. de interrup. Pouv. coupure 10kA 100kA flinke Keramik-Sicherung // Fusible cerámico rápido // Fusible céramique rapide – 38 – 2015.qxd 7/21/00 5:41 PM Page 41 Œ After disconnecting test leads and turning off the multimeter, remove the rear case by unscrewing the three screws that secure it to the front case. a - mA Input Jack Fuse: Remove the blown fuse (6.4 x 25.4mm) from the fuse holder. Replace with a 0.5A/250V quick acting fuse. Wavetek replacement fuse part number is FP 250 (a pack of four fuses). b - 20A Input Fuse: Remove the 20A fuse (10.3x38.1mm) from the 20 Ampere fuse holder. Replace with Wavetek FP425, Littel Fuse KLK-20 or Bussman KTK-20. c - Battery replacement: Remove battery and replace with NEDA type 1604, JIS006P, IEC 6F22 or equivalent 9-volt alkaline battery. Ž Reassemble the instrument. Warning: Use only an equivalent fuse to the one specified. Use of an incorrect fuse could result in serious injury or even death. Warning: Failure to turn off the multimeter before installing the battery could result in damage to the instrument and to the battery if the battery is connected incorrectly to the multimeter. D • Batterie/Sicherungsaustausch (siehe Fig 7) Warnung: Um elektrischen Schock zu vermeiden, Gerät vor Öffnen abschalten und Meßkabel entfernen Œ Geräterückseite losschrauben und abheben. a - mA Eingangssicherung: Defekte 6.4 x 25.4 mm Sicherung entfernen und mit 0.5A/250V flinker Sicherung ersetzen (Teil Nummer FP250; 4er Packung). b - 20A Eingangssicherung: Defekte 10.3 x 38.1 mm Sicherung entfernen und mit gleichwertiger 20A flinker Keramik Sicherung ersetzen (Wavetek FP425, Littel Fuse KLK-20 oder Bussman KTK-20). c Batterieaustausch: Batterie entfernen und durch gleichwertige 9V Batterie (NEDA 1604, JIS006P, IEC 6F22) ersetzen. Ž Geräterückseite wieder anbringen und festschrauben. Warnungen: Nur gleichwertige Sicherungen verwenden. Nicht-Abschalten des Gerätes bei Batterieaustausch kann Batterie und Gerät zerstören. E • Sustitución de la Pila y los Fusibles (vea fig. 7) Advertencia: Para evitar el peligro de descarga eléctrica, apague el multímetro y desconecte las puntas de prueba antes de abrir la tapa posterior. Œ Quite los tres tornillos y levante la tapa posterior. a Fusible de la entrada de mA: retire el fusible abierto (6.4 x 25.4 mm) y sustitúyalo por uno de actuación rápida y 0.5A/250V (ref. FP250, paquete con 4). b Fusible de la entrada de 20 A: retire el fusible abierto (10.3 x 38.1 mm) y sustitúyalo por otro cerámico, de actuación rápida y 20A (Wavetek FP425, Littel Fuse KLK-20 or Bussman KTK-20). c Sustitución de la pila: Retire la pila usada y cámbiela por otra de 9V de – 39 – 2015.qxd 7/21/00 5:41 PM Page 42 tipo NEDA 1604, JIS006P, IEC 6F22 o equivalente, preferiblemente alcalina. Ž Vuelva a cerrar el instrumento. Advertencia: Utilice solamente fusibles equivalentes a los especificados. Advertencia: Si no apaga el multímetro antes de cambiar la pila, podría causar daños al instrumento y a la pila F • Remplacement Pile et Batterie (voir fig 7) Avertissement: Pour éviter des chocs électriques et/ou des dégats à l’appareil, éteignez celui-ci et enlevez les cordons avant de l’ouvrir. Œ Dévissez le boîtier arrière et enlevez-le a - Fusible d’entrée mA: Enlevez le fusible défectueux 6.4 x 25.4 mm mm et remplacer le par un nouveau fusible rapide 0.5A/250V (Référence FP250; boîte de quatre). b - Fusible d’entrée 20A: Enlevez le fusible défectueux 10.3 x 38.1 mm Sicherung et remplacer le par un nouveau fusible céramique rapide 20A (Wavetek FP425, Littel Fuse KLK-20 ou Bussman KTK-20).  c Remplacement de la pile: Enlevez la pile, et remplacez la par une pile équivalente 9V (NEDA 1604, JIS006P, IEC 6F22). Ž Réassemblez l’appareil. Avertissements: Ne remplacez les fusibles que par des fusibles équivalents. Ne pas éteindre l’appareil pour le remplacement des piles peut endommager la plie et l’appareil. REPAIR Read the warranty located at the front of this manual before requesting warranty or non-warranty repairs. For warranty repairs, any multimeter claimed to be defective can be returned to any Wavetek Meterman authorized distributor or to a Wavetek Meterman Service Center for an over-the-counter exchange for the same or like product. Non-warranty repairs should be sent to a Wavetek Meterman Service Center. Please call Wavetek Meterman or enquire at your point of purchase for the nearest location and current repair rates. All multimeters returned for warranty or non-warranty repair or for calibration should be accompanied by the following information or items: company name, customer's name, address, telephone number, proof of purchase (warranty repairs), a brief description of the problem or the service requested, and the appropriate service charge (for non-warranty repairs). Please include the test leads with the meter. Service charges should be remitted in the form of a check, a money order, credit card with expiration date, or a purchase order made payable to Wavetek Meterman or to the specific service center. For minimum turn-around time on out-of-warranty repairs please phone in advance for service charge rates. The multimeter should be shipped with – 40 – 2015.qxd 7/21/00 5:41 PM Page 43 transportation charges prepaid to one of the following addresses or to a service center: in U.S.A. in Canada in Europe Wavetek Meterman 1420 75th Street SW Everett, WA 98203 Tel: 1-877-596-2680 Fax: 425-446-6390 Wavetek Meterman 400 Britannia Rd. E.Unit #1 Mississauga, ON L4Z 1X9 Tel: (905) 890-7600 Fax: (905) 890-6866 Wavetek Meterman 52 Hurricane Way Norwich, NR6 6JB, U.K. Tel:int + 44-1603-404824 Fax:int + 44-1603-482409 The instrument will be returned with the transportation charges paid by Wavetek Meterman. D • Reparatur Lesen Sie die Gewährl ei stung bevor Sie eine Reparat ur unter oder außerhalb Gewährleistung anfragen. Unter Gewährleistung bringen Sie bitte das def ekte Gerät zu ei ner aner kannten Wavetek Meterman Verkaufsstelle oder Servicestelle füreinen direkten Umtausch. Außerhalb Gewährlei st ung senden Sie das Gerät zu einer Wavetek Meterman anerkannten Servicestelle. Bitte informieren Sie sich bei Wavetek Meterman oder ihrem Fachhändler nach der dichtst beigelegen Adresse und nach aktuellen Reparaturgebühren. Bitte senden Sie folgende Informationen und Dokumente mit: Firmenname, Kundenname, Adresse, Telefoonnummer, Kaufnachweis (für Reparaturen unter Gewährl eistung), ei ne kurz e Beschr eibung der gewünschten Handlung, und di e gefordert e Bezahl ung ( Eingr iffe außerhal b der Gewährleistung). Bitte auch Testkabel beifügen. Bezahlungen in Form eines Checks, Bezahlungsformulieren, Kredietkarte mit Verf alldatum, usw. bitte in Namen der Servicestelle aufstellen. Bitte Multimeter (Frei) senden an: in U.S.A. in Canada in Europa Wavetek Meterman 1420 75th Street SW Everett, WA 98203 Tel: 1-877-596-2680 Fax: 425-446-6390 Wavetek Meterman 400 Britannia Rd. E.Unit #1 Mississauga, ON L4Z 1X9 Tel: (905) 890-7600 Fax: (905) 890-6866 Wavetek Meterman 52 Hurricane Way Norwich, NR6 6JB, U.K. Tel:int + 44-1603-404824 Fax:int + 44-1603-482409 oder an die Ihnen mitgetei lte Adresse. Multimeter wird (Frei) zurück geschickt E • Reparación Lea las condiciones de garantía, al principio de este manual, antes de solicitar cualquier reparación dentro o fuera de garantía. Si la reparación es en garantía, puede llevar el multímetro defectuoso a cualquier Distribuidor – 41 – 2015.qxd 7/21/00 5:41 PM Page 44 Autorizado o Centro de Servicio de Wavetek Meterman, donde le cambiarán en mano el producto por otro igual o similar. Para reparaciones fuera de garantía deberá enviar el multímetro a un Centro de Servicio de Wavetek Meterman. En Wavetek Meterman, o en su Distribuidor o punto de venta, le indicarán el Centro de Servicio más próximo y las tarifas de reparación vigentes. La documentación que acompañe a todo multímetro enviado para reparación debe incluir los siguientes datos: nombre de la empresa, persona de contacto, dirección, número de teléfono, prueba de compra (para reparaciones en garantía), una breve descripción del problema o el servicio requerido y, en caso de reparaciones fuera de garantía, si desea presupuesto previo. Por favor envíe las puntas de prueba con el multímetro. El importe de la reparación se enviará en forma de cheque, tarjeta de crédito con fecha de expiración u orden de pago a favor de Wavetek Meterman o del Centro de Servicio específico. El multímetro se enviará a portes pagados a una de las siguientes direcciones, o al Centro de Servicio que le hayan indicado: en EE. UU. en Canadá en Europa Wavetek Meterman 1420 75th Street SW Everett, WA 98203 Tel: 1-877-596-2680 Fax: 425-446-6390 Wavetek Meterman 400 Britannia Rd. E.Unit #1 Mississauga, ON L4Z 1X9 Tel: (905) 890-7600 Fax: (905) 890-6866 Wavetek Meterman 52 Hurricane Way Norwich, NR6 6JB, U.K. Tel:int + 44-1603-404824 Fax:int + 44-1603-482409 Wavetek Meterman devolverá el multímetro reparado a portes pagados. F • Réparation Lisez la garantie au début de ce manuel avant de demander une réparation sous garantie ou hors gar antie. Pour une réparation sous garantie , adressez-vous à votre revendeur Wavetek Meterman ou à un centre de services agréé par Wavetek Meterman pour un échange direct. Pour une réparation hors garantie, envoyez votre multimètre à un Centre de Services agrée par Wavetek Meterman . Téléphonez à Wavetek Meterman ou demandez à votre revendeur pour l'adresse la plus proche. Pour les réparations hors garantie, demandez dabord les tarifs. Joignez les informations et documents suivants: nom de sociètè, nom du client, adresse, numéro de téléphone, preuve d'achat (pour réparations sous garantie), une brève description de l'intervention souhaitée et le payement (pour réparations hors garantie). Ajoutez également les cordons de test. Le payement, sous forme de chèque, virement, carte de crédit avec date d'expiration, etc. doit êtr e éta bli au nom du Centre de Servic es. Le multimètre doit être envoyé port payé à: – 42 – 2015.qxd 7/21/00 5:41 PM Page 45 en U.S.A. en Canada en Europa Wavetek Meterman 1420 75th Street SW Everett, WA 98203 Tel: 1-877-596-2680 Fax: 425-446-6390 Wavetek Meterman 400 Britannia Rd. E.Unit #1 Mississauga, ON L4Z 1X9 Tel: (905) 890-7600 Fax: (905) 890-6866 Wavetek Meterman 52 Hurricane Way Norwich, NR6 6JB, U.K. Tel:int + 44-1603-404824 Fax:int + 44-1603-482409 ou à l'adresse communiquée. Le multimètre vous sera renvoyé port payé. – 43 – 2015.qxd 7/21/00 5:41 PM Page 46 ® T Manual Revision Date 07/00 Manual Part Number 1566149 Information contained in this manual is proprietary to Wavetek Meterman and is provided solely for instrument operation and maintenance. The information in this document may not be duplicated in any manner without the prior approval in writing from Wavetek. Meterman. Specifications subject to change. WAVETEK is a trademark of Wavetek Wandel Goltermann. © Wavetek Meterman, 2000 U.S. Service Center Wavetek Meterman 1420 75th Street SW Everett, WA 98203 Tel: (877) 596-2680 Fax: 425-446-6390 Canadian Service Center Wavetek Meterman 400 Britannia Rd. E.Unit #1 Mississauga, ON L4Z 1X9 Tel: (905) 890-7600 Fax: (905) 890-6866 European Distribution Center Wavetek Meterman 52 Hurricane Way Norwich, NR6 6JB, England Tel: (44) 1603-404-824 Fax: (44) 1603-482-409
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46

Wavetek 2015 Manual de usuario

Categoría
Multimetros
Tipo
Manual de usuario