Ohaus FC5916 120V Manual de usuario

Tipo
Manual de usuario
Frontier
TM
Centrifuge
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Instruction Manual
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Manual de instrucciones
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Manuel d’instructions
Frontier
TM
Zentrifuge
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Bedienungsanleitung
Frontier
TM
Centrifuga
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Manuale di Istruzioni
Sketch map of FC5714 / FC5718 / FC5816 / FC5916
Figure.1
Sketch map of FC5718R
Figure.2
Sketch map of FC5816R / FC5916R
Figure.3
1 Centrifuge Lid
2 Rotor Window
3 Display
4 Function Label
5 Main Power Switch
6 Power Connection
7 Emergency Release
Function Label for FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4
1
7
4
2
1
3
6
5
6
7
2
3
4
1
5
6
7
2
3
Figure.4
Figure.5
1. LCD Display
2. Program setup model
3. Start centrifugation
4. RPM/RCF model and select
5. Acceleration/Deceleration
intensity model and select
6. Temperature setup model
(Only FC5515R)
7. Time setup model
8. Store setup information
9. Short/quick spin centrifugation
10. Release lid
11. Stop centrifugation / setup
12. Adjusting knob/Dial: Change the
number
LCD Display
The following picture shows the individual elements of the LCD-display.
Figure.6
1
2
3
4
5
6
7
8
9
10
11
12
Figure.7
Display fields:
A-1 Display field "RPM/RCF"
A-2 Display field "Acc/Dec" "Service"
A-3 Display field "Time/Prog"
A-4 Display field –"Temp"
Messages/logos of the display fields
M1 "close"
M2 "open"
M3 "rotor"
M4 "Rotor-No."
M5 "rpm"
M6 "rcf"
M7 "accel"
M8 "decel"
M9 "radius"
M10 "program"
M11 "error"
M12 "service"
M13 "h m s"
M14 "temperature"
M15 "precool"
M14
M5
M6
M4
A-1
A-2
A-3
A-4
M2
M3
M1
M7
M8
M12
M15
M9
M11
M10
M13
Rotor Information Table
Rotor No.
display
Order No.
Capacity
Compatible
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
61
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
20
30314820
Rotor, Swing out, 4x250ml, ID
21
30314821
Rotor, Angle, 6x250ml FB, ID
22
30314822
Rotor, Swing out, 4x100ml, ID
23
30314823
Rotor, Swing out, 4x100ml, ID, Sealable
24
30314824
Rotor, Swing out, 2x3MTP, ID
25
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
26
30314826
Rotor, Angle, 6x85ml RB, ID
27
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
28
30314828
Rotor, Swing out, 16x50ml, ID
29
30314829
Rotor, Angle, 10x50ml FA, ID
30
30314830
Rotor, Angle, 6x50ml RB/FA, ID
31
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
32
30314832
Rotor, Angle, 30x15ml RB/FA, ID
33
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
34
30314834
Rotor, Angle, 12x15ml RB/FA, ID
35
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
36
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
38
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
39
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
40
30314840
Rotor, Angle, 64x0.5ml, ID
41
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
85
30553085
Rotor SwingOut 4x750ml D99mm ID SL
86
30553086
Rotor Angle 4x500ml D69mm ID
TABLE OF CONTENTS
1. INTRODUCTION ............................................................................................................................. 1
1.1 DESCRIPTION AND INTENDED PURPOSE...................................................................................... 1
1.2 BRIEF DESCRIPTION .................................................................................................................... 1
1.3 DEFINITION OF SIGNAL WARNINGS AND SYMBOLS....................................................................... 1
1.4 SAFETY PRECAUTIONS...................................................................................................................................................... 2
1.4.1 USER ................................................................................................................................................................................. 2
1.4.2 ROTOR AND ACCESSORIES .................................................................................................................................... 2
1.4.3 MEASURES FOR YOUR PROTECTION.................................................................................................................. 2
1.4.4 EXCLUDE THE FOLLOWING ENVIRONMENTAL INFLUENCES..................................................................... 2
1.4.5 MEASURES FOR OPERATIONAL SAFETY ........................................................................................................... 3
1.4.6 DANGER AND PRECAUTIONS.................................................................................................................................. 3
1.4.7 ABBREVIATIONS USED IN THIS MANUAL............................................................................................................ 3
2. INSTALLATION ......................................................................................................................................................................... 4
2.1 UNPACKING ............................................................................................................................................................................ 4
2.1.1 DELIVERY PACKAGE................................................................................................................................................... 4
2.2 SELECTING THE LOCATION ............................................................................................................................................. 4
2.3 INSTALLATION ....................................................................................................................................................................... 4
2.4 SAFETY PRECAUTIONS DURING OPERATION AND WARRANTY ....................................................................... 5
3. OPERATION .................................................................................................................................. 5
3.1 MOUNTING AND LOADING ROTOR ................................................................................................................................ 5
3.1.1 INSTALLATION OF ROTORS ..................................................................................................................................... 5
3.1.2 LOADING ANGLE ROTORS........................................................................................................................................ 6
3.1.3 LOADING SWING OUT ROTORS.............................................................................................................................. 6
3.1.4 LOADING AND OVERLOADING OF ROTORS ...................................................................................................... 7
3.1.5 REMOVING THE ROTOR ............................................................................................................................................ 7
3.2 LID CONTROL ......................................................................................................................................................................... 7
3.2.1 LID OPEN ......................................................................................................................................................................... 7
3.2.2 LID LOCK.......................................................................................................................................................................... 8
3.3 PRESELECTION ..................................................................................................................................................................... 8
3.3.1 PRESELECTION OF SPEED / RCF-VALUE ........................................................................................................... 8
3.3.2 PRESELECTION OF RUNNING TIME...................................................................................................................... 9
3.3.3 PRESELECTION OF BRAKE INTENSITY AND ACCELERATION.................................................................... 9
3.3.4 PRE-SELECTION OF TEMPERATURE (ONLY REFRIGERATED MODELS).............................................. 10
3.3.5 PRE-COOLING (ONLY REFRIGERATED MODELS).......................................................................................... 10
3.4 RADIUS CORRECTION ...................................................................................................................................................... 10
3.5 PROGRAM.............................................................................................................................................................................. 11
3.5.1 STORAGE OF PROGRAMS...................................................................................................................................... 11
3.5.2 RECALL OF STORED PROGRAMS ....................................................................................................................... 12
3.5.3 LEAVING PROGRAM MODE .................................................................................................................................... 12
3.6 STARTING AND STOPPING THE CENTRIFUGE........................................................................................................ 12
3.6.1 STARTING THE CENTRIFUGE................................................................................................................................ 12
3.6.2 THE "STOP" KEY ......................................................................................................................................................... 13
3.7 IMBALANCE DETECTION.................................................................................................................................................. 13
4. SETTING ................................................................................................................................................................................... 14
4.1 BASIC ADJUSTMENTS....................................................................................................................................................... 14
4.1.1 ACCESS TO MODE "OPERATING DATA" ............................................................................................................ 14
4.1.2 TEMPERATURE INDICATION .................................................................................................................................. 14
4.1.3 SIGNAL TURN ON / OFF ........................................................................................................................................... 15
4.1.4 VOLUME PRE-SELECTION OF SOUND SIGNAL............................................................................................... 15
4.1.5 SONG SELECTION FOR SOUND SIGNAL - END OF RUN ............................................................................. 16
4.1.6 KEYBOARD SOUND TURN ON / OFF ................................................................................................................... 16
4.1.7 CALL UP OPERATING DATA ................................................................................................................................... 16
5. MAINTENANCE....................................................................................................................................................................... 17
5.1 MAINTENANCE AND CLEANING .................................................................................................................................... 17
5.1.1 GENERAL...................................................................................................................................................................... 17
5.1.2 CLEANING AND DISINFECTION OF THE UNIT ................................................................................................. 18
5.1.3 CLEANING AND DISINFECTION OF THE ROTOR ............................................................................................ 18
5.1.4 DISINFECTION OF ALUMINUM ROTORS ............................................................................................................ 18
5.1.5 DISINFECTION OF PP-ROTORS ............................................................................................................................ 18
5.1.6 GLASS BREAKAGE..................................................................................................................................................... 19
5.2 LIFE TIME OF ROTORS, BUCKETS, ACCESSORIES............................................................................................... 19
6. TROUBLESHOOTING ........................................................................................................................................................... 19
6.1 ERROR MESSAGE: CAUSE / SOLUTION..................................................................................................................... 19
6.2 SURVEY OF POSSIBLE ERROR MESSAGES AND THEIR SOLUTIONS............................................................ 19
6.2.1 LID RELEASE DURING POWER FAILURE (EMERGENCY LID RELEASE)................................................ 19
6.2.2 DESCRIPTION OF THE ERROR MESSAGE SYSTEM ..................................................................................... 20
7. RECEIPT OF CENTRIFUGES TO REPAIR ..................................................................................................................... 21
8. TRANSPORT, STORAGE AND DISPOSAL..................................................................................................................... 21
8.1 TRANSPORT ......................................................................................................................................................................... 21
8.2 STORAGE............................................................................................................................................................................... 21
8.3 TRANSPORTING, INSTALLING, TRANSFERRING AND DISPOSING OF THE CENTRIFUGE
FC5714/FC5718/ FC5718R/FC5816/FC5816R .................................................................................................................... 21
8.3.1 TRANSPORT ................................................................................................................................................................. 21
8.3.2 INSTALLATION ............................................................................................................................................................. 21
8.3.3 PACKING........................................................................................................................................................................ 22
8.3.4 PASSING ON THE DEVICE ...................................................................................................................................... 22
9. TECHNICAL DATA ................................................................................................................................................................. 23
9.1 SPECIFICATIONS ................................................................................................................................................................ 23
9.1.1 CENTRIFUGE FC5714................................................................................................................................................ 23
9.1.2 CENTRIFUGE FC5718................................................................................................................................................ 24
9.1.3 CENTRIFUGE FC5718R............................................................................................................................................. 25
9.1.4 CENTRIFUGE FC5816................................................................................................................................................ 26
9.1.5 CENTRIFUGE FC5816R............................................................................................................................................. 27
9.2 DRAWINGS AND DIMENSIONS....................................................................................................................................... 30
10. COMPLIANCE ......................................................................................................................................................................... 31
11. APPENDIX ................................................................................................................................................................................ 33
11.1 TABLE 1: PERMISSIBLE NET WEIGHT....................................................................................................................... 34
11.2 TABLE 2: LOWEST TEMPERATURES AT MAX. SPEED (ONLY REFRIGERATED MODELS).................... 36
11.3 TABLE 3: MAX. SPEED AND RCF-VALUES FOR PERMISSIBLE ROTORS ..................................................... 37
11.4 TABLE 4: ACCELERATION AND DECELERATION TIMES .................................................................................... 40
11.5 TABLE 5: ERROR MESSAGES ...................................................................................................................................... 42
11.6 TABLE 6: RADIUS CORRECTION AND ADAPTER SPECIFICATIONS .............................................................. 43
11.7 TABLE 7: REDEMPTION FORM / DECONTAMINATION CERTIFICATE ............................................................ 47
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-1
1. INTRODUCTION
1.1 Description and Intended Purpose
Thank you for choosing this OHAUS product.
All symbols indicate safety instructions and points to potential dangerous situations. Please read the manual
completely before using the Frontier
TM
MultiPro Centrifuges to avoid incorrect operation.
Frontier
TM
MultiPro Centrifuges were designed for the separation of materials or mixtures with different densities.
OHAUS centrifuges are intended exclusively for indoor use and for use by qualified personnel.
1.2 Brief description
The models FC5714/FC5718/FC5816/FC5916 are non-refrigerated universal centrifuges.
The models FC5718R/FC5816R/FC5916R are refrigerated universal centrifuges.
All models are offered in two voltage variations, 230V or 120V.
The centrifuges can be used with swing-out and angle rotors.
All parameters are accessible via buttons and selected with the central adjuster. All pre-selected and current values
will be shown permanently on the LCD-display.
The centrifuge is powered by a maintenance-free induction motor.
Detailed technical data are in the "Technical data" section.
1.3 Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring
the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results.
The degree of danger is a part of a safety note and distinguishes the possible results of non-observance from each
other.
Signal Words
DANGER
Will lead to severe injuries or death if not avoided.
WARNING
For a hazardous situation with medium risk, possibly resulting in injuries or death if not
avoided.
CAUTION
For a hazardous situation with low risk, resulting in damage to the device or
the property or in loss of data, or injuries if not avoided.
ATTENTION
For important information about the product. May lead to equipment damage if not avoided
NOTE
For useful information about the product
Warning Symbols
General Hazard Electrical Shock Hazard
Alternating Current
Biohazard
Explosion Crushing
EN-2 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Warning and information signs on the surface of centrifuge
Four carriers must be used at all times on four place swing out rotors or damage will
occur to the centrifuge. Such damage will not be covered under the product warranty.
Attention! Check the fastening of the rotor nut before each run.
Take off mains plug before opening the housing or the emergency release.
Direction of rotation clockwise rotation for the rotor drive
Reference for loading rotors
1.4 Safety Precautions
1.4.1 User
OHAUS centrifuges are intended exclusively for indoor use and for use by qualified personnel. This device may
only be operated by trained specialist staff. They must have carefully read the operating manual and be familiar
with the functions of the device.
1.4.2 Rotor and accessories
Only OHAUS original rotors and accessories shall be used. Any other use or intended use is considered improper.
OHAUS is not liable for damage resulting from improper use.
CAUTION:
Read all safety warnings before installing, making connections, or servicing this equipment. Failure to
comply with these warnings could result in personal injury and/or property
damage. Retain instructions for future reference.
1.4.3 Measures for your protection
WARNING: Never work in an environment subject to explosion hazards! The housing of the instrument is
not gas tight. (Explosion hazard due to spark formation, corrosion caused by the ingress of gases)
WARNING: When using chemicals and solvents, comply with the instructions of the producer and the
general lab safety rules.
WARNING: The centrifuge is not sealed. Use suitable protection measures when using the centrifuge for
infectious and pathogenic samples. Follow appropriate safety precautions when handling these samples.
1.4.4 Exclude the following environmental influences
Powerful vibrations
Direct sunlight
Atmospheric humidity greater than 80%
Corrosive gases present
Temperatures below 2 °C and above 35 °C
Powerful electric or magnetic fields:
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-3
WARNING:
Electrical shock hazards exist within the housing. The housing should only be opened by authorized
and qualified personnel. Remove all power connections to the unit before opening.
1.4.5 Measures for operational safety
Do not unscrew the two halves of the housing
Dry off any liquid spills immediately! The instrument is not watertight
Verify that the equipment’s input voltage range and plug type are compatible with the local power supply.
Only connect the power cord to a properly grounded power receptacle.
Only use a power cord with a rating that exceeds the specifications on the equipment label.
Do not position the equipment such that it is difficult to disconnect the power cord from the power receptacle.
Make sure that the power cord does not pose a potential obstacle or tripping hazard.
The equipment is for indoor use only. Use the equipment only in dry locations.
Use only approved accessories.
Operate the equipment only under ambient conditions specified in these instructions.
Disconnect the equipment from the power supply when cleaning.
Do not operate the equipment in hazardous or unstable environments.
Service should only be performed by authorized personnel.
1.4.6 Danger and precautions
To protect people and environment the following precautions should be observed:
During centrifugation, the presences of people and the setting up of hazardous materials are prohibited within
30 cm around the centrifuge according to the regulations of EN 61010-2-020.
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R are not explosion-proof and must therefore not
be operated in explosion-endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or
such substances, which chemically react with high energy, is strictly prohibited. The final decision on the risks
associated with the use of such substances is the responsibility of the user of the centrifuge.
Never spin toxic or pathogenic material without adequate safety precautions, i.e. centrifugation of buckets /
tubes with missing or defective hermetic sealing is strictly prohibited. The user is obliged to perform appropriate
disinfection procedures in case dangerous substances have contaminated the centrifuge and or its accessories.
When centrifuging infectious substances, always pay attention to the general laboratory precautions. If
necessary, contact your safety officer!
It is prohibited to run the centrifuge with rotors other than listed for this unit.
Under no circumstances open the lid of the centrifuge while the rotor is still running or rotating with a speed of >
2m/s
1.4.7 Abbreviations used in this manual
Symbol/Abbreviations
Unit
Description
RPM
[min
-1
]
rpm
revolutions per minute
RCF
[x g]
relative centrifugal force
PCR
Polymerase chain reaction
PP
-
Polypropylene
PC
-
Polycarbonate
accel
-
acceleration
decel
-
deceleration
prog
-
program
EN-4 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
2. INSTALLATION
2.1 Unpacking
Carefully remove your centrifuge and each of its components from the package. The included components vary
depending on the centrifuge model (see table below). Save the packaging to ensure safe storage and transport.
The instruction manual must always be kept with the centrifuge!
Rotor(s) / Accessories will be packed separately.
WARNING: Lifting Hazard. Single person lift could cause injury. Use a mechanical lifting
device or team lifting procedures when lifting or moving the equipment.
Please refer to section 8.3 for details about lifting the equipment out of the packaging.
2.1.1 Delivery package
2.2 Selecting the Location
NOTE!
Avoid excessive vibrations, heat sources, air current, or rapid temperature changes.
The centrifuge should be installed on an even, solid and level surface, if possible on a laboratory cabinet / table
or some other solid vibration free surface.
During centrifugation, the centrifuge must be placed in a way, that there is a minimum space of 30 cm on each
side of the unit according to the standards EN 61010-2-020.
Do not place the centrifuge next to a window or a heater, where it could be exposed to excessive heat, as the
performance of the unit is based on an ambient temperature of 2C.
2.3 Installation
Follow these steps:
Check whether the power supply corresponds with the one specified on the manufacturer's rating label, which
is located on the rear panel.
For FC5714/FC5718/FC5816, the power line should be protected by a 10 A rating circuit breaker (type K).
For FC5718R/FC5816R /FC5916/FC5916R, the power line should be protected by a 16 A rating circuit breaker
(type K).
In case of emergency, there must be an emergency switch off installed outside the room in order to disconnect
the power supply from the unit.
Connect the centrifuge to a grounded power receptacle.
Connect the centrifuge with the mains. (The socket for the power cord must be easy to reach for disconnection)
Turn the instrument on using the mains power switch.
Open the lid by using the Door Open button.
Remove the transport securing device of the motor.
Quantity
Description
1
Centrifuge
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
1
Power Cable
1
Warranty Card
1
Instruction Manual/Quick Guide
1
Rotor Key
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-5
2.4 Safety precautions during operation and warranty
Do not operate the centrifuge in case it is not installed correctly.
Do not lean on the centrifuge during operation.
Do not stay within the 30 cm clearance envelope longer than necessary for operational reasons.
Do not place any potentially hazardous materials within the 30 cm clearance envelope.
Do not operate the centrifuge when disassembled (e.g. without housing).
Do not run the centrifuge when mechanical or electrical components have been tampered with.
Do not use accessories such as rotors and buckets, which are not exclusively approved by OHAUS
Corporation, except commercially available centrifuge tubes made of glass or plastic.
Do not spin extremely corrosive substances, as they may damage or weaken the materials.
Do not operate the centrifuge with rotors or buckets, which show any signs of corrosion or mechanical damage.
The manufacturer is responsible for safety and reliability of the centrifuge, only if:
1) The unit is operated in accordance with this instruction manual.
2) Modifications, repairs or other adjustments are performed by OHAUS authorized personnel and the
electrical installation complies with the relevant electrical code.
NOTE!
Warranty
The centrifuge has been subjected to thorough testing and quality controls. In the unlikely case of any
manufacturing faults occurring, the centrifuge and rotors are covered by warranty. This warranty becomes
invalid in case of mishandling, damage and negligence and further in case of usage of inappropriate spare parts
and / or accessories or unauthorized modification of the unit.
Technical modification rights are reserved by the manufacturer in respect to technical improvement!
3. OPERATION
3.1 Mounting and loading rotor
3.1.1 Installation of rotors
Clean the drive shaft as well as the collet with a clean, grease-free piece of cloth. Place the rotor onto the drive
shaft. (See figure below). Take care that the rotor is fully installed onto the motor shaft.
Figure.8
Figure. 9 Figure.10
Nut for Rotor
Tool for rotor
with nut
Tool for rotor
without nut
Snap-on lid
Screw-on lid
Motor shaft and chamber
EN-6 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Hold the rotor with one hand and secure the rotor to the shaft by turning the fixing nut clockwise. Tighten the fixing
nut with enclosed rotor key (See figures 9-10)
We will provide a tool for none-nut rotor with centrifuge, the tool for nut-rotor will be provided with rotor.
ATTENTION!
Check that the fixing screw is properly installed before each run. (See figure 9-10)
Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or mechanical
damage.
Do not operate with extremely corrosive substances, which could damage the rotor, buckets and
materials.
In case of any questions, please contact the manufacturer!
3.1.2 Loading angle rotors
Rotors must be loaded symmetrically and with equal weight (See figure below). The adapter may only be loaded
with the appropriate vessels. The weight differences between the filled vessels should be kept as low as possible.
Therefore we recommend weighing them with a balance. This reduces the wear of the drive and the acoustic
operating noise.
The maximum load per hole is stated on each rotor.
Figure.11 WRONG Figure.12 CORRECT ( 6 tubes)
3.1.3 Loading swing out rotors
Loading of the buckets / vessels must be made in accordance to the figure below.
It is allowed to operate e.g. a 4-place-rotor with 2 loaded buckets only. But the loaded buckets must be opposite to
each other. Make sure that the unloaded buckets also be put inside the rotor (see below).
In principle swing out rotors may not be taken into operation until all buckets or racks are put into the rotor.
The bolts at the rotor must be greased with the "High TEF oil". The sample tubes have to be filled evenly by eye
and put into the drillings or tube racks. The weight difference of the loaded buckets should not exceed approx.1.0
g.
ATTENTION!
Swing out rotors may be taken in operation only if all locations are filled in with either four buckets
or four carriers – do not mix buckets and carriers together!!
ATTENTION!
Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or
mechanical damage.
Do not operate with extremely corrosive substances, which could damage the rotor and buckets.
In case of any questions, please contact the manufacturer!
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-7
Figure.13 WRONG Figure.14 CORRECT
3.1.4 Loading and overloading of rotors
All approved rotors are listed with their maximum speed and maximum filling weight in "table 2 permissible net
weight" (See APPENDIX).
The maximum load permitted for a rotor, which is determined by the manufacturer, as well as the maximum speed
allowed for this rotor (See label on rotor), must not be exceeded. The liquids the rotors are loaded with should have
a maximum homogeneous density of 1.2 g/ml or less when the rotor is running at maximum speed.
In order to spin liquids with a higher density, the speed has to be reduced according to the following formula:
Reduced speed n
red
=
1,2
𝑖𝑔𝑒𝑟 𝑑𝑒𝑛𝑠𝑖𝑡𝑦
x max. speed (n
max
) of the rotor
Example:
n
red
=
1,2
1,7
x 4.000 = 3.360 rpm
If in case of any questions, please contact the manufacturer!
To determine the relative centrifugal force(RCF/G-force) for a specific adapter, you can calculate per DIN 58 970
using the attached formula:
RCF = 1.117862*10
5
*n
2
*r
max
n: revolutions per minute (RPM)
r
max
: max centrifuging radius in cm by using the bottom of tubes
3.1.5 Removing the rotor
Untighten the rotor fixing nut completely (screw over the stiff point) and lift the rotor vertically out of the centrifuge.
(See figure 9 and 10).
3.2 Lid control
3.2.1 Lid open
After the run, when the lid of the centrifuge is closed, the word "close" (M1) appears in the display "RPM | RCF"
(A-1). Additionally, if there is a rotor in the centrifuge, the word "rotor"(M3) appears as well as the code number of
the respective rotor, which is in the centrifuge system "71" (M4). If there is no rotor in the centrifuge, the word
"rotor" (M3) flashes and additionally the word "no" (M4) appears. By pressing the key "Door Open" (7) you can
release the lid of the centrifuge. As soon as the electromagnetic lid is completely released, the word "open" (M2)
appears. Now you can open the lid of the centrifuge.
Please refer to figure 15 below for reference.
EN-8 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figure. 15
During the run you can call up the rotor type at any time by pressing the key "Door Open" (7).
3.2.2 Lid lock
The lid should only be put down slightly. An electromagnetic lid lock closes the lid, at the same time the word
"open" (M2) disappears (refer to figure 15).
As a sign that the centrifuge is ready for starting, in the display "RPM | RCF" (A-1) the word "close" (M1)
appears. Simultaneously the word "rotor" (M3) is displayed, as well as the code number of the rotor, which is in
the centrifuge system, "no 71" (M4). With that, all rotor specific data, like e. g. max. speed, acceleration etc., are
adopted.
ATTENTION:
Don't grip your fingers between lid and device or locking mechanism when closing the lid!
3.3 Preselection
3.3.1 Preselection of speed / RCF-value
This pre-selection is activated through the key "RPM | RCF" (4) (refer to figure 16 below). By pressing the key
once the word "rpm" (M5) flashes. By pressing the key twice the pre-selection of the centrifugal forces can be
selected. Then the flashing word "rcf" (M6) appears. You can set the desired values with the adjusting knob (1). In
the display (A-1) the regulated value is shown permanently, before, during and after the run.
Figure. 16
As long as no rotor is inserted, the speed is adjustable between 200 rpm and maximum revolution of the centrifuge.
If there is a rotor in the centrifuge the speed can only be pre-selected until the maximum permissible revolution of
that rotor. It is the same with the pre-selection of the RCF-value. The setting range is between the centrifugal force
at 200 rpm and the maximum permissible centrifugal force of the rotor.
See "Table 3: max. speed and RCF-values for permissible rotor" (See APPENDIX). All important values are
listed there.
M5
A-1
4
5
M6
M3
A-1
M2
M4
M1
7
1
1
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-9
ATTENTION:
Please also check the maximum permissible revolutions of your test tubes with the manufacturer.
3.3.2 Preselection of running time
The running time can be pre-selected in three different ranges from 10 seconds up to 99 hours 59 minutes.
1. Range from 10 seconds up to 59 minutes 50 seconds in steps of 10 seconds
2. Range from 1 hour up to 99 hours 59 minutes in steps of 1 minute.
3. The continuous run "cont", which can be interrupted by the key "Stop"(10) (refer to figure 17).
The running time can be pre-selected with the lid open or closed.
To activate the setting of the running time press the key "Time" (6).
In the display "Time/Prog" (A-3) flashes the indication "m : s" or "h : m", depending on the previous setting.
To set the desired value, use the adjusting knob (1). After exceeding 59 min 50 sec the indication changes
automatically into "h : m". After exceeding 99 hours 59 min the word "cont" appears in the display "Time/Prog"
(A-3). That continuous run can only be interrupted by pressing the key "Stop" (10). The time countdown starts as
soon as the set speed is reached.
The display always shows the remaining running time. (See figure 17)
Figure. 17
3.3.3 Preselection of brake intensity and acceleration
This function is activated through the key "Accel/Decel" (5) (refer to figure18).
By pressing the key once the word "accel" (M7) flashes in the display "Acc/Dec" (A-2). The desired acceleration
can be pre-selected by the adjusting knob (1). The value 0 is equivalent to the lowest and the value 9 to the highest
acceleration.
By pressing the key "Accel/Decel" (5) twice, the display "Acc/Dec" (A-2) indicates the word "decel"(M8). Now
the desired brake intensity can be pre-selected by the adjusting knob (1). The value 9 is equivalent to the shortest
and the value 0 to longest possible brake time.
See "table 5: acceleration and deceleration times" (APPENDIX). There the acceleration and deceleration times
for the acceleration and deceleration stages 0 to 9 for permissible rotors are shown.
Figure. 18
6
A-3
1
10
M7
M8
5
A-2
1
EN-10 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
3.3.4 Pre-selection of temperature (Only Refrigerated Models)
This function is activated by the key "Temp/Setup" (13). After pressing this key in the display "Time/Prog" the
indication "°C " (A-4) flashes. By the adjusting knob (1) the desired test temperature can be pre-selected in steps of
C in a range from -20°C up to +40°C.
The value is indicated permanently in the display (figure 19) - before, during and after the run.
Please notice the respective lowest temperatures of the rotors at maximum speed!
Figure. 19
3.3.5 Pre-cooling (Only Refrigerated Models)
If the samples are temperature-sensitive it is useful to pre-cool the centrifuge, the rotor and eventually the buckets
to the required working temperature. Therefore, insert the desired rotor and pre-set the respective temperature. By
simultaneous pressing the keys "Temp/Setup" (13) (refer to figure 20) and "Time" (6) you start the run. While
running, the unit chooses automatically a rotational speed that is equivalent to 20 % of the permitted rotational
speed of the respective rotor. After the pre-set temperature is reached you can leave the pre-cooling run with the
"Stop" key (10).
Depending on the inserted rotor the pre-cooling goes between approx. 10 and 20 min.
Figure. 20
3.4 Radius correction
If you use adapters or reducers it could change the centrifugal radius of the respective rotor. In that case you can
correct the radius manually. Please proceed as follows:
Close the lid, then press the key "Time" (6) (refer to figure 21) and the key "Prog/Setup" (11) at the same time
and hold them.
In the display "Time/Prog" (A-3) the word "radius" (M9) appears. By the adjusting knob (1) you can preselect the
respective radius correction (See Table 7, APPENDIX) in steps of 0.1 cm. As soon as you have set a radius
correction the word "radius" (M9) appears. This word will be visible until you put the radius correction back to 0
again.
A-4
1
13
A-4
13
6
10
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-11
Figure. 21
3.5 Program
3.5.1 Storage of programs
You can store up to 99 runs with all relevant parameters, including the used rotors. You can use any free program
number and call it up again.
Put the needed rotor into the centrifuge. By pressing the key "Prog/Setup" (11) in the display "Time/Prog" (A-3)
the word "programm" appears. With the adjusting knob (1) you can chose the desired program number.
If a program number is already occupied, in the display "RPM | RCF" (A-1), the words "rotor" (M3) and "xx" (M4)
will appear. In case of free program numbers, 0 appears.
Figure. 22
Close the lid of the centrifuge. Now proceed as described previously to set all important run parameters. If the lid
isn't closed when storing the program, in the display "RPM/RCF" (A-1), the words "FirSt" and "CLOSE Lid" (See
figure 23) flashes alternately. If you want to start the run without storing the program, in the display "RPM/RCF" (A-
1) ,the words "First" and "PrESS StoreE" (See figure 24) flashes alternately.
Figure. 23 Figure. 24
11
A-3
A-1
M3
M4
M10
1
12
1
11
A-3
M9
6
EN-12 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
For adaption of data press the key "Store" (12) (refer to figures 23 and 24) for approx. 1 second. If the program is
stored correctly, the word "StorE" appears in the display "RPM/RCF" (A-1). As a result, the word "programm"
(M10) disappears.
As soon as the key "Store" (12) is released, the word "programm xx" (M10) reappears the (xx) stands for the
chosen program location.
If all program numbers are occupied you can take an old number that is not necessary anymore and just put in the
new parameters.
3.5.2 Recall of stored programs
To recall stored programs press the key "Prog/Setup" (11) (refer to figure 25) while the lid is already closed. Inside
the display "Time/Prog" (A-3), "programm –"(M10) appears. The desired program number can be pre-selected
with the adjusting knob (1).
In the respective displays the stored values for that program will appear.
If the wrong rotor is inside the centrifuge for the pre-selected program, in the display "RPM | RCF" (A-1), the word
"rotor" (M3) flashes. At the same time the word "FALSE" and the stored rotor number " xx" (M4) will flashing by
turns.
Figure. 25
3.5.3 Leaving program mode
To leave the program mode just press the key "Prog/Setup" (11) (refer to figure 25). Then inside the display
"Time/Prog" the word "programm" appears.
Set the display to "programm--" (M10) with the adjusting knob (1).
3.6 Starting and stopping the centrifuge
3.6.1 Starting the centrifuge
You can start the centrifuge either with the "Start" key (9) (refer to figure 26) or the "Quick Spin" key (8).
By the "Start" key (9) you can start stored runs or runs with manually pre-selected parameters.
When the respective pre-selected running time has ended the centrifuge will stop automatically.
By the "Quick Spin" key (8) you can start runs, which will last just a few seconds.
By pressing the "Quick Spin" key (8) the centrifuge accelerates up to the pre-selected revolution.
In the display "Time/Prog" (A-3) the passed running time is indicated from the date of pressing the "Quick Spin"
key (8).
By releasing the "Quick Spin" key (8) the centrifuge stops and the running time is indicated until the opening of
the lid.
A-3
M10
A-1
M3
M4
1
11
12
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-13
Figure. 26
3.6.2 The "STOP" key
By the "Stop" key (10) (See figure 27) you can interrupt the run at any time. After pressing the key the centrifuge
decelerates with the respective pre-selected intensity down to stand still.
Figure. 27
3.7 Imbalance detection
In case of the rotor not being equally loaded, the drive will turn off during acceleration. The rotor decelerates to
stand still.
When in the display "Time/Prog" (A-3) the word "error” (M11) together with the number "01” appear, the weight
difference of the samples is too large. Distribute the weight evenly.
Load the rotor as described in chapter 3.1.2 and 3.1.3.
When inside the display "Time/Prog" (A-3) the word "error” together with the number "02" (See figure 28) appear,
it could be due to the following reason: The imbalance switch is defective.
Figure. 28
10
A-3
M11
1
8
9
A-3
EN-14 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4. SETTING
4.1 Basic adjustments
4.1.1 Access to mode "Operating Data"
When using the centrifuge, the following parameters can be set:
- Temperature indication °C or °F
- Acoustic signal turn on/off
- Keyboard sound turn on/off
- Volume pre-selection of sound signal
- Song selection of sound signal "end of run"
While the centrifuge is turned off, press simultaneously the keys "Time"(6) and "Door Open" (7) and turn on the
main switch of the centrifuge. Now release both keys and as a result a display test is executed for approx. 5
seconds. All indicators will appear at the same time (See figure 29).
Figure. 29
ATTENTION:
Please notice that you must enter the program as described under point 4.1.1 to change the
adjustments of the points 4.1.2 - 4.1.7. After you have stored the settings you can change to normal
program mode again by switching off the centrifuge for a short while.
All changed settings must be confirmed by the key "Start"(9). The word "Store"(12) appears in the display "RPM |
RCF"(A-1) - Only then the pre-selections are valid!
Figure. 30
4.1.2 Temperature indication
Proceed as described under point 4.1.1 to enter this program mode and then press the key "Accel/Decel" (5). In
the display "Acc/Dec" (A-2) the word "Service" appears. Now select the letter "C" with the adjusting knob (1). As
a result, in the display "RPM | RCF" (A-1), the words "CELSI/temp" appear. If you press the key "RPM | RCF" (4),
6
7
A-1
9
12
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-15
the word "CELSI" flashes and you can change the display into Fahrenheit "FAREN", with the adjusting knob (1)
(See figure 31).
After you have stored the settings (See 4.1.1) you change back to the normal program mode again by switching off
the centrifuge for a short while.
Figure. 31
4.1.3 Signal turn on / off
Proceed as described under point 4.1.1 to enter this program mode and then press the key "Accel/Decel" (5). In
the display "Acc/Dec" (A-2) the word "Service" flashes. Now select the letter "L" with the adjusting knob (1). As a
result, the words "On Sound" appears in the display "RPM | RCF" (4). If you press the key "RPM | RCF" (4) now,
the word "On" flashes and you can switch off the sound with the adjusting knob (1) (See figure 32).
After you have stored the settings (See 4.1.1) you change back to the normal program mode again by switching off
the centrifuge for a short while.
Figure. 32
4.1.4 Volume pre-selection of sound signal
Proceed as described under point 4.1.1to enter this program mode and then press the key "Accel/Decel" (5). In
the display "Acc/Dec" (A-2) the word "Service" flashes. Now select the letter "U" with the adjusting knob (1). As a
result, in the display "RPM | RCF" (A-1) the words "Vol=0- 9/Sound" appear. After pressing the key "RPM | RCF"
(4), you can adjust the desired volume between 0 (low) and 9 (loud) with the adjusting knob (1) (See figure 33).
After you have stored the settings (see 4.1.1) you can change back to the normal program mode again by
switching off the centrifuge for a short period.
Figure. 33
4
5
1
4
5
1
A-2
A-1
4
5
1
A-2
A-1
1
EN-16 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4.1.5 Song selection for sound signal - end of run
Proceed as described under point 4.1.1to enter this program mode and then press the key "Accel/Decel" (5). In
the display "Acc/Dec" (A-2) the word "Service" flashes. Now select the letter "G". with the adjusting knob (1). As
a result, in the display "RPM | RCF" (A-1), the word "SonGo/Sound" appears. After pressing the key "RPM |
RCF" (4), you can select a song with the adjusting knob (1). (See figure 34).
After you have stored the settings (see 4.1.1) you can change back to the normal program mode again by switch
off the centrifuge for a short while.
Figure. 34
4.1.6 Keyboard sound turn on / off
Proceed as described under point 4.1.1to enter this program mode and then press the key "Accel/Decel" (5). In
the display "Acc/Dec" (A-2) the word "Service" flashes. Now select the letter "B". with the adjusting knob (1). As
a result, in the display "RPM | RCF" (A-1), the word "ON/BEEP " appears. After pressing the key "RPM | RCF"
(4), you can turn the keyboard sound (On) or (Off) with the adjusting knob (1). (See figure 39).
After you have stored the settings (see 4.1.1) you can change back to the normal program mode again by switch
off the centrifuge for a short while.
Figure. 35
4.1.7 Call up operating data
ATTENTION:
This should only be performed by advanced user or service engineer.
In the mode "Basic Adjustments" you can call up the operating data of the centrifuge. Please proceed as
described under point 4.1.2 to enter this program mode. Press the key "Accel/Decel" (5). In the display "Acc/Dec"
(A-2) the word "Service" flashes.
With the adjusting knob (1) the different information can be accessed:
A= previous starts of the centrifuge
4
5
1
A-2
A-1
4
5
1
A-2
A-1
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-17
H= previous operating hours
S= software version
r= converter software
E= list of previous error messages
h= running time of the motor
The list of the last 99 error messages can be looked over by pressing the key "RPM | RCF" (4) and scroll through it
by the adjusting knob (1). The respective error codes appear in the display "RPM | RCF" (A-1). Please refer to
"Table 6: error messages" (see APPENDIX).
To change back to normal program mode again, switch off the centrifuge for a short period.
Figure. 36
5. MAINTENANCE
5.1 Maintenance and cleaning
5.1.1 General
Care:
Maintenance of the centrifuge is confined to keeping the rotor, the rotor chamber and the rotor accessories clean
as well as to regularly lubricating the rotor insert bolts of a swing out rotor (if available).
The most suitable lubricant is the High TEF oil.
Lubricants containing molycote and graphite are not allowed.
Please pay special attention to anodized aluminum parts. Breakage of rotors can be caused even by slight
damage.
In case of rotors, buckets or tube racks getting in touch with corrosive substances the concerned spots have to be
cleaned carefully.
Corrosive substances are for instance: alkalis, alkaline soap solutions, alkaline amines, concentrated acids,
solutions containing heavy metals, water-free chlorinated solvents and saline solutions. e.g. salt water, phenol,
halogenated hydrocarbons.
Cleaning – units, rotors, accessories:
Turn the device off and disconnect it from the power supply before you begin any cleaning or disinfecting. Do
not pour liquids into the housing interior.
Do not spray disinfectant on the device.
Thorough cleaning not only has its purpose in hygiene but also in avoiding corrosion due to pollution.
In order to avoid damaging anodized parts such as rotors, reduction plates etc., only pH-neutral Detergents
with a pH-value of 6-8 may be used for cleaning. Alkaline cleaning agents (pH-value > 8) must not be used.
4
5
1
A-2
A-1
EN-18 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
After cleaning, please ensure all parts are dried thoroughly, either by hand or in a hot-air cabinet
(max.Temperature + 50°C).
It is necessary to coat anodized aluminum parts with anti-corrosion oil regularly in order to increase their life-
spans and reduce corrosion predisposition.
Due to humidity or not hermetically sealed samples, condensate may be formed. The condensate has to be
removed from the rotor chamber with a soft cloth regularly.
The maintenance procedure has to be repeated every 10 to 15 runs, or at least once a week.
- Connect the unit to the power supply, after the equipment is completely dry.
- Do not carry out disinfection with UV-, beta- and gamma-rays or other high energy radiation.
- Metal rotors can be autoclaved.
- Rotor lid and adapters can also be autoclaved (max. 12C, 20 min).
- The tube racks are made of PP and cannot be autoclaved at 134°C.
5.1.2 Cleaning and disinfection of the unit
1. Open the lid before you turn off the unit. Disconnect it from the power supply.
2. Open the rotor nut by turning the rotor key counter clockwise.
3. Remove the rotor.
4. For cleaning and disinfection of the unit and the rotor chamber use the above mentioned cleaner.
5. Clean all accessible areas of the device and its accessories, including the power cord with a damp cloth.
6. Wash the rubber seals and rotor chamber thoroughly with water.
7. Rub the dry rubber seals with glycerol or talc to prevent these to becoming brittle. Other components of the
unit, e.g. the lid lock, motor shaft and rotor must not be greased.
8. Dry the motor shaft with a soft, dry and lint-free cloth.
9. Control the unit and accessories for damage.
Make sure that the centrifuge is turned off the unit and disconnect the unit from the power supply. Then remove
adherent dust from the ventilation slots in the centrifuge by using a soft brush. Do this at least every six months.
5.1.3 Cleaning and disinfection of the rotor
1. Clean and disinfect the rotors, rotor lids and adapters with the above mentioned cleaner.
2. Use a bottle brush to clean and disinfect the rotor bores.
3. Rinse the rotors, rotor lid and adapter with clear water. Particularly the drillings of angle rotors.
4. For drying of the rotors and accessories set them on a towel. Place the angle rotors with bores down.
5. Dry the rotor cone with a soft, dry and lint-free cloth and look for damage. Do not grease the rotor cone.
5.1.4 Disinfection of aluminum rotors
In case of infectious material spilling into the centrifuge, the rotor and rotor chamber have to be disinfected directly
after the run. Rotors may be autoclaved at a maximum temperature of 121°C.
5.1.5 Disinfection of PP-rotors
Autoclaving
The recommended time for autoclaving: 15 – 20 min at 121°C (1 bar)
ATTENTION:
The sterilization time of 20 min. must not be exceeded. Repeated sterilization will cause
reduction of the mechanical resistance of the plastic material
Before autoclaving the PP-rotor and adapter must be thoroughly cleaned to avoid the burning in of dirty residues.
You can disregard the consequences of some chemical residues to plastic materials at ambient temperatures. But
at the high temperatures during autoclaving those residues may corrode and destroy the plastic. The objects must
be thoroughly rinsed with distilled water after the cleaning but before the autoclaving. Residues of any cleaning
liquids may cause fissures, whitening and stains.
Gas sterilization
Adapters, bottles and rotors may be gas sterilized with Ethylenoxyd. Make sure to air out the items after the
sterilization and before using them again.
ATTENTION:
Because the temperature may rise during the sterilization, rotors, adapters and bottles must not
be closed and must be totally unscrewed
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-19
Chemical sterilization
Bottles, adapters and rotors may be treated with the usual liquid disinfectants.
ATTENTION:
Before applying any other cleaning or decontamination method than recommended by the
manufacturer, contact the manufacturer to ensure that it will not damage the unit or the rotor.
5.1.6 Glass breakage
With high g-values, the rate of glass tube breakage increases. Glass splinters have to be removed immediately
from rotor, buckets, adapters and the rotor chamber itself. Fine glass splinters will scratch and therefore damage
the protective surface coating of a rotor. If glass splinters remain in the rotor chamber, fine metal dust will build up
due to air circulation. This very fine, black metal dust will significantly pollute the rotor chamber, the rotor, the
buckets and the samples.
If necessary, replace the adapters, tubes and accessories to avoid further damage. Check the rotor bores regularly
for residues and damage.
ATTENTION:
Please check the relevant specifications of the tubes centrifuges with the manufacturer.
5.2 Life time of rotors, buckets, accessories
Rotors and rotor lid made of aluminum or stainless steel, have an operating time of max. 7 years from first use.
Transparent rotor lids and caps made of PC or PP as well as rotors, tube racks and adapters of PP have a
maximum operating time up to 3 years from first use.
Condition for the operating time: Proper use damage-free condition, recommended care.
6. TROUBLESHOOTING
6.1 Error message: Cause / Solution
The error messages are listed to help localize possible errors faster.
The diagnosing referred to in this chapter may not always be the case, as they are only theoretically occurring
errors and solutions.
Please keep us informed about any kind of error occurring, which is not listed in this chapter. Only through your
information are we able to improve this operation manual.
Many thanks in advance for your support.
6.2 Survey of possible error messages and their solutions
6.2.1 Lid release during power failure (Emergency Lid Release)
In case of power failure or malfunction, the lid of the centrifuge can be opened manually in order to retrieve your
samples.
For FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (motor driven lock)
Please proceed as follows:
1. Switch off the centrifuge and unplug the power cord, wait until the rotor has come to a standstill (this may take
several minutes)
2. On the left side of the centrifuge housing there is a plastic stopper. Remove this stopper and behind it there is a
hexagon nut.
3. Take the included box spanner, put it in the hole and lock the box spanner with the hexagon nut (See figure
37).
4. Now turn the box spanner to the right side (clockwise) up to the limit.
ATTENTION: 3
a) Just turn to the limit, don´t tighten the nut.
b) Now open the lid of the centrifuge.
c) Switch the centrifuge on again, to resume work.
EN-20 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figure. 37
For 5714
Please proceed as follows (see Figure 38):
ATTENTION:
Switch the centrifuge off and unplug the power cord, wait until the rotor stands still (this may
take several minutes). At the right side of the centrifuge there is a plastic stopper (Figure 38).
Remove this stopper, which is connected to the lid lock, horizontally from the housing until the
centrifuge lid opens.
Now open the lid of the centrifuge
Figure. 38
6.2.2 Description of the error message system
The error message "error" (M11) is shown in the "Time/Prog" (A-3) display (See figure 39). Detailed information
about possible error messages are in "table 6: error messages" (See Appendix).
Figure. 39
A-3
M11
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-21
7. RECEIPT OF CENTRIFUGES TO REPAIR
Health risk from contaminated equipment, rotors and accessories.
In case of returning the centrifuge for repairing to the manufacturer, please notice the following:
The centrifuge must be decontaminated and cleaned before the shipment for the
protection of persons, environment and material.
Decontamination certificate at goods return delivery (See APPENDIX)
We reserve the right to not accept contaminated centrifuges.
Further on all costs occurred for the cleaning and disinfection of the units will go to
the debit of the customer’s account.
8. TRANSPORT, STORAGE AND DISPOSAL
8.1 Transport
Before transporting, take out the rotor.
Only transport the unit in the original packaging.
Install the transport protection material to secure the motor shaft, when
transporting over longer distances.
Air temperature rel. humidity Air pressure
General transportation -25 to 60 °C 10 to 75 % 30 to 106 kPa
8.2 Storage
During storage of the centrifuge the following environmental conditions must
be observed:
Air temperature rel. Humidity Air pressure
in transport packaging -25 to 55 °C 10 to 75 % 70 to 106 kPa
8.3 Transporting, Installing, Transferring and Disposing of the Centrifuge FC5714/FC5718/
FC5718R/FC5816/FC5816R
These instructions complement the previous general instructions in chapter 8 and do not replace them.
8.3.1 Transport
Please transport the device in the original packaging.
The centrifuge should always be transported using a mechanical
transport device.
8.3.2 Installation
Opening the carton and lifting out the device.
1. Cut the adhesive tape.
2. Open all 4 flaps of the carton.
3. Remove the accessories.
4. Carefully lift the centrifuge from the carton.
WARNING: Lifting Hazard. Single person lift could cause injury. Use a mechanical lifting
device or team lifting procedures when lifting or moving the equipment.
EN-22 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Place the device on a stable, horizontal and non-resonant lab bench
1. Remove the front and back transport protection material.
2. Remove the plastic sleeve.
3. Observe a minimum distance of 30 cm to adjoining devices at the sides and
from the rear side to the wall.
4. Install the device in a well-ventilated location which is protected from direct
sunlight to prevent it from overheating.
Connect the device
1. After installation, wait for four hours before switching the centrifuge on in order to avoid damage to the
compressor.
2. Check that the mains voltage and frequency match the requirements on the device name plate(see rear side of
the device) and then connect the device to the power supply.
Remove the transport protection material from the rotor chamber
1. Switch on the device at the mains power switch.
2. Open the centrifuge lid using the open button.
3. Remove the transport protection material.
4. Place the rotor vertically onto the motor shaft.
5. Turn the rotor nut using the rotor key clockwise until the rotor nut is tightened.
The device is now ready to use
Retain the packaging and all transport protection material for shipping the device at a later date.
8.3.3 Packing
Pack the centrifuge in reverse order.
8.3.4 Passing on the Device
When passing the equipment on to third parties, please make sure to also include this instruction manual.
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-23
9. TECHNICAL DATA
9.1 Specifications
9.1.1 Centrifuge FC5714
Model
FC5714
Speed Range
200 rpm -14000 rpm;10 rpm/set
Maximum RCF
18624 x g;10 x g/set
Maximum Capacity(Rotor)
4x100ml
Temperature range(N/A)
Air cool
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise level (depending on the rotor)
63 ± 2 dB(A)
Allowable density at maximum speed
1.2 g/ml
Allowable kinetic energy
5595 Nm
Mains power connection AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
1.3 A
2.4 A
Power consumption
240 W
300 W
Dimensions (W × D × H)
362 x 493 x 330 mm
14.3 x 19.4 x 13.0 in
Net Weight (without rotor)
30 kg
66 lb
Shipping Dimensions (W × D × H)
580 x 490 x 460 mm
22.8 x 19.3 x 18.1 in
Shipping Weight (without rotor)
32.5 kg
72 lb
Ambient conditions (EN/IEC 61010-1)
Environment
For indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Overvoltage category (IEC 60364-4-443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
EN-24 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.2 Centrifuge FC5718
Model
FC5718
Speed Range
200 rpm -18000 rpm;10 rpm/set
Maximum RCF
23542 x g;10 x g/set
Maximum Capacity (Rotor)
4x100ml
Temperature range (N/A)
Air cool
Running Time Noise level
10 sec to 99 hr 59 min 59 sec or continuous
Noise level (depending on the rotor)
60 ± 2 dB(A)
Allowable density at maximum speed
1.2 g/ml
Allowable kinetic energy
16672 Nm
Mains power connection AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
2.0 A
4.0 A
Power consumption
455 W
475 W
Dimensions (W × D × H)
408 x 499 x 351 mm
16.1 x 19.7 x 13.8 in
Net Weight (without rotor)
43 kg
95 lb
Shipping Dimensions (W × D × H)
650 x 520 x 490 mm
25.6 x 20.5 x 19.3 in
Shipping Weight (without rotor)
53 kg
117 lb
Ambient conditions (EN/IEC 61010-1)
Environment
for indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Overvoltage category (IEC 60364-4-443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-25
9.1.3 Centrifuge FC5718R
Model
FC5718R
Speed Range
200 rpm -18000 rpm;10 rpm/set
Maximum RCF
23542 x g;10 x g/set
Maximum Capacity(Rotor)
4x100ml
Temperature range(Digital)
-20~40
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise level (depending on the rotor)
60 ± 2 dB(A)
Allowable density at maximum speed
1.2 g/ml
Allowable kinetic energy
25111 Nm
Mains power connection AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
3.0 A
6.0 A
Power consumption
660 W
660 W
Dimensions (W × D × H)
407 x 731 x 359 mm
16.0 x 28.8 x 14.1 in
Net Weight (without rotor)
60 kg
132 lb
Shipping Dimensions (W × D × H)
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
Shipping Weight (without rotor)
77 kg
170 lb
Ambient conditions (EN/IEC 61010-1)
Environment
for indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Overvoltage category (IEC 60364-4-443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
EN-26 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.4 Centrifuge FC5816
Model
FC5816
Speed Range
200 rpm -15000 rpm;10 rpm/set
Maximum RCF
21379 x g;10 x g/set
Maximum Capacity(Rotor)
6 x 250ml
Temperature range(N/A)
Air cool
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise level (depending on the rotor)
61 ± 2 dB(A)
Allowable density at maximum speed
1.2 g/ml
Allowable kinetic energy
34363 Nm
Mains power connection AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
2.4 A
4.2 A
Power consumption
530 W
520 W
Dimensions (W × D × H)
446 x 538 x 354 mm
17.6 x 21.2 x 13.9 in
Net Weight (without rotor)
52 kg
115 lb
Shipping Dimensions (W × D × H)
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
Shipping Weight (without rotor)
77 kg
170 lb
Ambient conditions (EN/IEC 61010-1)
Environment
for indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Overvoltage category (IEC 60364-4-443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-27
9.1.5 Centrifuge FC5816R
Model
FC5816R
Speed Range
200 rpm -16000 rpm;10 rpm/set
Maximum RCF
24325 x g;10 x g/set
Maximum Capacity(Rotor)
6 x 250ml
Temperature range(Digital)
-20~40
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise level (depending on the rotor)
63 ± 2 dB(A)
Allowable density at maximum speed
1.2 g/ml
Allowable kinetic energy
34363 Nm
Mains power connection AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
3.7 A
7.8 A
Power consumption
785 W
850 W
Dimensions (W × D × H)
723 x 538 x 354 mm
28.5 x 21.2 x 13.9 in
Net Weight (without rotor)
77 kg
170 lb
Shipping Dimensions (W × D × H)
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
Shipping Weight (without rotor)
87 kg
192 lb
Ambient conditions (EN/IEC 61010-1)
Environment
for indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Overvoltage category (IEC 60364-4-443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
EN-28 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.6 Centrifuge FC5916
Model
FC5916
Speed Range
200 rpm -16000 rpm;10 rpm/set
Maximum RCF
24325 x g;10 x g/set
Maximum Capacity(Rotor)
4x750ml
Temperature range(N/A)
Air cool
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise level (depending on the rotor)
63 ± 2 dB(A)
Allowable density at maximum speed
1.2 g/ml
Allowable kinetic energy
60629 Nm
Mains power connection AC
230 V ~
50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
2.8 A
5.6 A
Power consumption
640 W
680 W
Dimensions (W × D × H)
540 x 670 x 390 mm
21.3 x 26.4 x 15.4 in
Net Weight (without rotor)
85 kg
187 lb
Shipping Dimensions (W × D × H)
780 x 670 x 590 mm
30.7 x 26.4 x 23.3 in
Shipping Weight (without rotor)
98 kg
216 lb
Ambient conditions (EN/IEC 61010-1)
Environment
For indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Overvoltage category (IEC 60364-4-443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-29
9.1.7 Centrifuge FC5916R
Model
FC5916R
Speed Range
200 rpm -16000 rpm;10 rpm/set
Maximum RCF
26331 x g;10 x g/set
Maximum Capacity(Rotor)
4x750ml
Temperature range(Digital)
-20~40
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise level (depending on the rotor)
63 ± 2 dB(A)
Allowable density at maximum speed
1.2 g/ml
Allowable kinetic energy
54458 Nm
Mains power connection AC
230 V ~
50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
7.2 A
20 A
Power consumption
1630 W
1750 W
Dimensions (W × D × H)
730 x 670 x 390 mm
28.7 x 26.4 x 15.4 in
Net Weight (without rotor)
118 kg
260 lb
Shipping Dimensions (W × D × H)
900 x 750 x 560 mm
40.0 x 29.5 x 22.0 in
Shipping Weight (without rotor)
137 kg
137 kg
Ambient conditions (EN/IEC 61010-1)
Environment
for indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Overvoltage category (IEC 60364-4-443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
EN-30 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.2 Drawings and dimensions
D
H
W
W
H
D
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-31
Model
W (mm / in.)
D (mm / in.)
H (mm / in.)
FC5714
362 / 14.3
493 / 19.4
330 / 13.0
FC5718
408 / 16.1
499 / 19.7
351 / 13.8
FC5718R
408 / 16.1
731 / 28.8
359 / 14.1
FC5816
446 / 17.6
538 / 21.2
354 / 13.9
FC5816R
723 / 28.5
538 / 21.2
354 / 13.9
FC5916
540 / 21.3
670 / 26.4
390 / 15.4
FC5916R
730 / 28.7
670 / 26.4
390 / 15.4
10. COMPLIANCE
Compliance to the following standards is indicated by the corresponding mark on the product.
Marking
Standard
This product complies with the applicable harmonized standards of EU Directives
2011/65/EU (RoHS), 2014/30/EU (EMC) and 2014/35/EU (LVD) and 2014/31/EU
(NAWI). The EU Declaration of Conformity is available online at
www.ohaus.com/ce.
Disposal
In conformance with the European Directive 2012/19/EU on Waste Electrical and
Electronic Equipment (WEEE) this device may not be disposed of in domestic
waste. This also applies to countries outside the EU, per their specific
requirements.
Please dispose of this product in accordance with local regulations at the
collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the
distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use),
the content of this regulation must also be related.
For disposal instructions in Europe, refer to www.ohaus.com/weee.
Thank you for your contribution to environmental protection.
EN-32 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
FCC Supplier Declaration of Conformity
Unintentional Radiator per 47CFR Part B
Trade Name: OHAUS CORPORATION
Model: FC...
Party issuing Supplier’s Declaration of Conformity:
Ohaus Instruments (Changzhou) Co., Ltd.
2F, 22 Block, 538 West Hehai Road, Xinbei District, Changzhou
Jiangsu 213022
China
Phone: +86 519 85287270
Responsible Party U.S. Contact Information:
Ohaus Corporation
7 Campus Drive, Suite 310
Parsippany, NJ 07054
United States
Phone: +1 973 377 9000
Web: www.ohaus.com
FCC Compliance Statement:
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference
in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of the following measures:
—Reorient or relocate the receiving antenna.
—Increase the separation between the equipment and receiver.
—Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
—Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s
authority to operate the equipment.
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-33
11. APPENDIX
TABLE 1: PERMISSIBLE NET WEIGHT
TABLE 2: LOWEST TEMPERATURES AT MAX. SPEED
TABLE 3: MAX. SPEED AND RCF-VALUES FOR PERMISSIBLE ROTORS
TABLE 4: ACCELERATION AND DECELERATION TIMES
TABLE 5: ERROR MESSAGES
TABLE 6 (PART 1): RADIUS CORRECTION
TABLE 7: REDEMPTION FORM / DECONTAMINATION CERTIFICATE
EN-34 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.1 Table 1: Permissible net weight
Rotor number
Rotor Description
Permissible net
weight
Max Speed ( x1000 rpm)
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
16.0
16.0
30314820
Rotor, Sw ing out, 4x250ml, ID
4x 536 g
4.5
4.5
30314821
Rotor, Angle, 6x250ml FB, ID
6x 355 g
8.0
8.0
8.0
8.0
30314822
Rotor, Sw ing out, 4x100ml, ID
4x 340 g
4.5
4.5
4.5
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
4x 465 g
4.0
5.0
5.0
30314824
Rotor, Sw ing out, 2x3MTP, ID
2x 310 g
4.5
4.5
4.5
4.5
4.5
4.5
4.5
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
6x 140 g
11.0
13.5
30314826
Rotor, Angle, 6x85ml RB, ID
6x 140 g
9.0
9.0
11.0
13.0
11.0
11.0
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
4x 140 g
12.0
12.0
12.0
12.0
15.0
15.0
30314828
Rotor, Sw ing out, 16x50ml, ID
4x 557 g
4.5
4.5
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-35
30314829
Rotor, Angle, 10x50ml FA, ID
10x 74 g
7.5
7.5
9.0
10.5
10.0
10.0
30314830
Rotor, Angle, 6x50ml RB/FA, ID
6x 72 g
6.0
6.0
6.0
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
6x 94 g
12.0
12.0
13.0
13.0
13.0
13.0
30314832
Rotor, Angle, 30x15ml RB/FA, ID
30x 32 g
4.5
4.5
4.5
4.5
4.5
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
20x 18 g
12.0
12.0
12.0
12.0
12.0
12.0
30314834
Rotor, Angle, 12x15ml RB/FA, ID
12x 25 g
6.0
6.0
6.0
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
44x 3.8 g
13.5
13.5
15.0
16.0
16.0
16.0
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
30x 3.4 g
12.0
13.0
14.0
15.0
15.0
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
15.0
15.0
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
12x 3.4 g
18.0
18.0
30314840
Rotor, Angle, 64x 0.5ml, ID
64x 2.3 g
13.5
13.5
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
4x 3.5 g
15.0
15.0
15.0
15.0
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
4x 995 g
4.0
4.0
30553086
Rotor Angle 4x 500ml D69mm ID
4x 708 g
8.0
8.0
EN-36 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.2 Table 2: Lowest temperatures at max. speed (Only Refrigerated Models)
Rotor number
Rotor Description
Model
Max Speed
N-max
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314821
Rotor, Angle, 6x250ml FB, ID
FC5916R
8000 rpm
-1 ° C
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5718R
4500 rpm
-7 ° C
FC5816R
4500 rpm
-7 ° C
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5718R
5000 rpm
-3 ° C
FC5816R
5000 rpm
-3 ° C
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5718R
4500 rpm
-6 ° C
FC5816R
4500 rpm
-6 ° C
FC5916R
4500 rpm
-8 ° C
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718R
13500 rpm
17 ° C
FC5816R
13500 rpm
17 ° C
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718R
9000 rpm
-5 ° C
FC5816R
9000 rpm
-5 ° C
FC5916R
13000 rpm
8 ° C
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718R
12000 rpm
5 ° C
FC5816R
12000 rpm
5 ° C
FC5916R
16000 rpm
7 ° C
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718R
7500 rpm
9 ° C
FC5816R
7500 rpm
9 ° C
FC5916R
10500 rpm
2 ° C
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5718R
6000 rpm
-8 ° C
FC5816R
6000 rpm
-8 ° C
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718R
12000 rpm
-5 °C
FC5816R
12000 rpm
-5 ° C
FC5916R
13000 rpm
-4 ° C
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5718R
4500 rpm
-9 ° C
FC5816R
4500 rpm
-9 ° C
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718R
12000 rpm
2 ° C
FC5816R
12000 rpm
2 ° C
FC5916R
12000 rpm
-2 ° C
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5718R
6000 rpm
-11.5 ° C
FC5816R
6000 rpm
-11.5 ° C
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718R
13500 rpm
1 ° C
FC5816R
13500 rpm
1 ° C
FC5916R
16000 rpm
3 ° C
30314836
Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable
FC5718R
14000 rpm
9 ° C
FC5816R
14000 rpm
9 ° C
FC5916R
15000 rpm
3 ° C
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718R
18000 rpm
-2 ° C
FC5816R
18000 rpm
-2 ° C
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718R
13500 rpm
2 ° C
FC5816R
13500 rpm
2 ° C
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718R
15000 rpm
-1 ° C
FC5816R
15000 rpm
-1 ° C
FC5916R
15000 rpm
-2 ° C
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916R
4500 rpm
4 ° C
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916R
8000 rpm
0 ° C
All temperature indications refer to a room temperature of 23°C. By exceeding this value or direct solar
radiation to the centrifuge, these values can´t be kept up.
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-37
11.3 Table 3: Max. speed and RCF-values for permissible rotors
Rotor number
Rotor Description
Model
Max Speed
RPM
RCF
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
4500
3780 x g
FC5816R
4500
3780 x g
30314821
Rotor, Angle, 6x 250ml FB, ID
FC5816
8000
10016 x g
FC5816R
8000
10016 x g
FC5916
8000
10016 x g
FC5916R
8000
10016 x g
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5714
4500
3350 x g
FC5718
4500
3350 x g
FC5718R
4500
3350 x g
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
4000
2486 x g
FC5718
5000
3885 x g
FC5718R
5000
3885 x g
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5714
4500
2716 x g
FC5718
4500
2716 x g
FC5718R
4500
2716 x g
FC5816
4500
2716 x g
FC5816R
4500
2716 x g
FC5916
4500
2716 x g
FC5916R
4500
2716 x g
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
11000
13932 x g
FC5718R
13500
20984 x g
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
9000
10413 x g
FC5718R
9000
10413 x g
FC5816
11000
15555 x g
FC5816R
13000
21726 x g
FC5916
11000
15555 x g
FC5916R
13000
21726 x g
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718
12000
14809 x g
FC5718R
12000
14809 x g
FC5816
12000
14809 x g
FC5816R
12000
14809 x g
FC5916
15000
23140 x g
FC5916R
16000
26328 x g
30314828
Rotor, Sw ing out, 16x 50ml, ID
FC5816
4500
3735 x g
FC5816R
4500
3735 x g
EN-38 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718
7500
8174 x g
FC5718R
7500
8174 x g
FC5816
9000
11771 x g
FC5816R
10500
16022 x g
FC5916
10000
14532 x g
FC5916R
10500
16022 x g
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
12000
13522 x g
FC5718R
12000
13522 x g
FC5816
13000
15869 x g
FC5816R
13000
15869 x g
FC5916
13000
15869 x g
FC5916R
13000
15869 x g
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5714
4500
2830 x g
FC5718
4500
2830 x g
FC5718R
4500
2830 x g
FC5816
4500
2830 x g
FC5816R
4500
2830 x g
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
12000
15775 x g
FC5718R
12000
15775 x g
FC5816
12000
15775 x g
FC5816R
12000
15775 x g
FC5916
12000
15775 x g
FC5916R
12000
15775 x g
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718
13500
17113 x g
FC5718R
13500
17113 x g
FC5816
15000
21128 x g
FC5816R
16000
24039 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5714
12000
15131 x g
FC5718
13000
17758 x g
FC5718R
14000
20595 x g
FC5916
15000
23643 x g
FC5916R
15000
23643 x g
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-39
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718
18000
23542 x g
FC5718R
18000
23542 x g
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
13500
16298 x g
FC5718R
13500
16298 x g
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
15000
15343 x g
FC5718R
15000
15343 x g
FC5916
15000
15343 x g
FC5916R
15000
15343 x g
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916
4000
3434 x g
FC5916R
4500
4346 x g
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916
8000
10367 x g
FC5916R
8000
10367 x g
EN-40 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.4 Table 4: Acceleration and deceleration times
Item No
Rotor Description
Model
Acceleration* Time in
sec
Deceleration* Time in
sec
level 0
level 9
level 0
level 9
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5714
200
25
230
35
FC5718
240
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
20
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
309
34
458
36
FC5816R
309
34
458
36
30314821
Rotor, Angle, 6x250ml FB, ID
FC5816
664
130
2906
92
FC5816R
664
130
2906
92
FC5916
573
66
1903
84
FC5916R
573
66
1903
84
30314822
Rotor, Sw ing out, 4x 100ml, ID
FC5714
110
14
170
19
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
110
14
170
17
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5714
230
25
340
26
FC5718
160
20
360
15
FC5718R
160
20
360
15
FC5816
452
43
616
38
FC5816R
452
43
616
38
FC5916
249
27
488
23
FC5916R
249
27
488
23
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
400
40
960
40
FC5718R
500
60
1260
50
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
360
40
1050
40
FC5718R
360
40
1050
40
FC5816
697
85
2313
70
FC5816R
2313
70
1630
76
FC5916
463
48
1654
46
FC5916R
549
69
1307
67
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718
300
60
820
40
FC5718R
300
60
820
40
FC5816
506
60
1745
49
FC5816R
506
60
1745
49
FC5916
448
50
1251
45
FC5916R
448
50
1251
45
30314828
Rotor, Sw ing out, 16x50ml, ID
FC5816
311
34
387
36
FC5816R
311
34
387
36
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718
380
63
796
37
FC5718R
380
62
1069
36
FC5816
753
115
2395
72
FC5816R
753
115
2395
72
FC5916
480
60
1747
68
FC5916R
480
60
1747
68
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5714
101
11
206
14
FC5718
89
13
239
11
FC5718R
90
12
463
11
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-41
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
446
48
1323
49
FC5816R
446
48
1323
49
FC5916
264
28
921
32
FC5916R
264
28
921
32
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5714
157
19
370
20
FC5718
160
15
380
10
FC5718R
160
15
380
10
FC5816
149
25
985
20
FC5816R
149
25
985
20
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
512
54
1439
47
FC5816R
512
54
1439
47
FC5916
305
32
988
37
FC5916R
305
32
988
37
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
102
11
167
14
FC5718
88
13
222
11
FC5718R
88
12
433
11
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718
196
22
204
18
FC5718R
196
22
204
18
FC5816
257
29
436
28
FC5816R
274
31
485
29
FC5916
233
25
283
30
FC5916R
233
25
500
25
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5714
250
27
280
34
FC5718
200
25
360
25
FC5718R
210
25
360
30
FC5916
221
23
561
30
FC5916R
221
23
561
30
30314838
Rotor, Angle, 24x 1.5/2.0ml, ID
FC5714
200
22
230
35
FC5718
210
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
30
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718
240
25
210
20
FC5718R
240
25
210
20
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
150
17
170
12
FC5718R
150
17
170
12
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
130
16
130
12
FC5718R
130
16
130
12
FC5916
100
12
201
12
FC5916R
100
12
201
12
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916
483
47
1287
49
FC5916R
551
55
1501
54
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916
575
73
2317
82
FC5916R
575
73
2317
82
*Note: accelerates from 0 to Vmax; decelerates from Vmax to 0.
EN-42 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.5 Table 5: Error messages
Error-No.: Description
1 Imbalance arose
2 Imbalance sensor is defective
4 Imbalance switch has been activated for longer than 5 seconds
8 Transponder in the rotor is defective
11 Temperature sensor is defective
12 Chamber over temperature
14
CLOSE lid
33 Open lid while motor is running
34 Lid contact defective
38 Lid motor is blocked
40 Communication with frequency converter distrubed during start
41 Communication with frequency converter distrubed during stop
42 Short circuit in the frequency converter
43 Undervoltage frequency converter
44 Overvoltage frequency converter
45 Over temperature frequency converter
46 Over temperature motor
47 Over current frequency converter
48 Timeout between control unit and frequency converter
49 Other error frequency converter
55 Overspeed
70 Timeout between controler and RS232 interface
99 Rotor is not allowed in this centrifuge
FALSE Inserted rotor does not exist in the programm
rotor no Rotor is not detected
Leap of speed is too big between two mesaurements
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-43
11.6 Table 6: Radius correction and adapter specifications
Rotor
Rotor
Description
Bucket/Ad
apter
Bucket/Adapter Description
Radius
(cm)
Correction
(cm)
Compatible
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle,
24x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314820
Rotor, Sw ing out,
4x 250ml, ID
Blank
NO Bucket/Adapter
16.7
30314903
Rack, 1x250ml D62mm FB, 2/pk
15.7
-1.0
30314904
Rack, 1x100ml D41mm RB, 2/pk
15.9
-0.8
30314905
Rack, 3x50ml D29mm RB, 2/pk
16.3
-0.4
30314906
Rack, 3x50ml D29mm FA, 2/pk
16.7
0.0
30314907
Rack, 1x50ml D34mm RB, 2/pk
15.8
-0.9
30314908
Rack, 2x 50ml D29mm Rim, 2/pk
15.8
-0.9
30314909
Rack, 1x50ml D29mm Rim, 2/pk
16.1
-0.6
30314910
Rack, 5x25ml D24.5mm RB, 2/pk
16.1
-0.6
30314911
Rack, 3x25ml D24.5mm Rim, 2/pk
15.5
-1.2
30314912
Rack, 9x15ml D17mm RB, 2/pk
16.3
-0.4
30314913
Rack, 7x15ml D17mm FA, 2/pk
16.4
-0.3
30314914
Rack, 7x15ml D17mm Rim, 2/pk
16.1
-0.6
30314915
Rack, 8x10ml D16mm RB, 2/pk
16.3
-0.4
30314916
Rack, 14x5-7ml D13mm RB, 2/pk
15.9
-0.8
30314917
Rack, 10x1.6-7ml D13mm Vac, 2/pk
15.9
-0.8
30304367
Rack, 8x4-10ml D16mm Vac, 2/pk
16.3
-0.4
30304368
Adapter, 9x 1.5/2.0ml D11mm, 2/pk
15.7
-1.0
30314821
Rotor, Angle,
6x 250ml FB, ID
Blank
NO Bucket/Adapter
14.1
30304369
Adapter, 8x 1.5ml D11mm, 2/pk
13.2
-0.9
30304370
Adapter, 5x 10ml D16mm RB, 2/pk
13.3
-0.8
30304371
Adapter, 4x 15ml D17mm FA, 2/pk
13.0
-1.1
30304372
Adapter, 2x 30ml D26mm RB, 2/pk
12.5
-1.6
30304373
Adapter, 1x 50ml D29mm RB, 2/pk
12.0
-2.1
30304374
Adapter, 1x 50ml D30mm FA, 2/pk
11.7
-2.4
30559412
Adapter 15x 1.5/2.0ml D11mm
MTBx 2 V1
9.5
13.6
-4.6
-0.5
30559413
Adapter 1x 250ml D60mm CBCx 3 V1
11.8
-0.5
30559414
Adapter 1x 175/225ml D61mm FAx 2
V1
12.8
-2.3
30314822
Rotor, Sw ing out,
4x 100ml, ID
Blank
NO Bucket/Adapter
14.8
30314842
Rack, 1x100ml D46mm RB, 2/pk
14.1
-0.7
30314843
Rack, 1x85ml D38mm RB, 2/pk
14.1
-0.7
30314844
Rack, 1x50ml D34mm RB, 2/pk
14.2
-0.6
30314845
Rack, 1x50ml D29mm RB, 2/pk
14.2
-0.6
30314846
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.7
30314847
Rack, 1x50ml D29mm Rim, 2/pk
14.3
-0.5
30314848
Rack, 1x30ml D25mm Rim, 2/pk
13.8
-1.0
30314849
Rack, 1x25ml D25mm RB, 2/pk
14.4
-0.4
30314850
Rack, 2x15ml D17mm FA, 2/pk
14.8
0.0
30314851
Rack, 1x15ml D17mm Rim, 2/pk
14.5
-0.3
EN-44 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314852
Rack, 7x15ml D17mm RB, 2/pk
14.5
-0.3
30314853
Rack, 9x5-7ml D13mm RB, 2/pk
14.2
-0.6
30314854
Rack, 10x1.5ml D11mm, 2/pk
9.9
-4.9
30314855
Rack, 9x4.5ml D12mm FB, 2/pk
14.0
-0.8
30314856
Rack, 7x5-7ml D13mm RB, 2/pk
11.6
-3.2
30314857
Rack, 7x5-7ml D13mm Vac, 2/pk
14.2
-0.6
30314858
Rack, 5x4-10ml D16mm Sar, 2/pk
14.3
-0.5
30314823
Rotor, Sw ing out,
4x 100ml, ID,
Sealable
Blank
NO Bucket/Adapter
14.6
30314860
Bucket, 100ml wo Cap, Sealable,
2/pk
14.2
-0.4
30314861
Bucket, 100ml w/ Cap, Sealable,
2/pk
14.2
-0.4
30314864
Rack, 1x100ml D40mm RB, 2/pk
13.8
-0.8
30314865
Rack, 1x85ml D38mm RB, 2/pk
14.0
-0.6
30314866
Rack, 1x50ml D34mm RB, 2/pk
14.0
-0.6
30314867
Rack, 1x50ml D29mm RB, 2/pk
14.0
-0.6
30314868
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.5
30314869
Rack, 1x50ml D29mm Rim, 2/pk
13.9
-0.7
30314870
Rack, 1x30ml D25mm Rim, 2/pk
13.1
-1.5
30314871
Rack, 1x25ml D25mm RB, 2/pk
14.0
-0.6
30314872
Rack, 1x15ml D17mm FA, 2/pk
14.1
-0.5
30314873
Rack, 2x15ml D17mm FA, 2/pk
14.1
-0.5
30314874
Rack, 2x15ml D17mm Rim, 2/pk
14.0
-0.6
30314875
Rack, 4x 15ml D17mm RB, 2/pk
14.0
-0.6
30314876
Rack, 7x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314877
Rack, 5x1.5ml D11mm, 2/pk
14.0
-0.6
30314878
Rack, 3x10ml D17mm Sar, 2/pk
14.0
-0.6
30314879
Rack, 4x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314880
Rack, 4x10ml D17mm Vac, 2/pk
14.0
-0.6
30314881
Bucket, 2x50ml D29mm FA, w /
Rack, 2/pk
14.6
0.0
30314882
Bucket, 3x15ml D17mm FA, w /
Rack, 2/pk
14.6
0.0
30314883
Cy to, Rectangular Bucket, 2/pk
6.3
-8.3
30314824
Rotor, Sw ing out,
2x 3MTP, ID
Blank
NO Bucket/Adapter
12.0
30314891
Adapter, 48x 1.5/2.0ml D11mm, 2/pk
12.0
0.0
30314825
Rotor, Angle,
6x 85ml RB, ID, Hi
Blank
NO Bucket/Adapter
10.3
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
9.6
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
9.6
-0.7
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.0
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
9.8
-0.5
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
9.3
-1.0
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
10.3
0.0
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
9.5
-0.8
30314826
Rotor, Angle,
6x 85ml RB, ID
Blank
NO Bucket/Adapter
11.5
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.9
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
10.4
-1.1
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.9
-0.6
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
10.6
-0.9
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
10.4
-1.1
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
11.1
-0.4
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
10.4
-1.1
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-45
30314827
Rotor, Angle,
4x 85ml RB, ID, Hi
Blank
NO Bucket/Adapter
9.2
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
8.5
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
8.4
-0.8
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
8.9
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
8.6
-0.6
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
8.3
-0.9
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
7.5
-1.7
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
8.3
-0.9
30314828
Rotor, Sw ing out,
16x 50ml, ID
Blank
NO Bucket/Adapter
16.5
30304375
Rack, 4x50ml D29mm FA, 2/pk
16.5
0.0
30314583
Rack, 10x15ml D17mm FA, 2/pk
16.5
0.0
30314829
Rotor, Angle,
10x 50ml FA, ID
Blank
NO Bucket/Adapter
13.0
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
12.8
-0.2
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.4
-2.6
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
8.9
-4.1
30314830
Rotor, Angle,
6x 50ml RB/FA, ID
Blank
NO Bucket/Adapter
11.0
30130891
Adapter, 1x 30ml D26mm, 2/pk
10.7
-0.3
30130892
Adapter, 1x 16ml D18mm, 2/pk
10.3
-0.7
30130893
Adapter, 1x 15ml D17mm, 2/pk
10.6
-0.4
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.4
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
8.3
-2.7
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
6.7
-4.3
30314831
Rotor, Angle,
6x 50ml RB, ID, Hi
Blank
NO Bucket/Adapter
8.4
30130891
Adapter, 1x 30ml D26mm, 2/pk
8.2
-0.2
30130892
Adapter, 1x 16ml D18mm, 2/pk
7.9
-0.5
30130893
Adapter, 1x 15ml D17mm, 2/pk
8.0
-0.4
30314892
Adapter, 1x 15ml D17.5mm FA, 2/pk
7.7
-0.7
30314832
Rotor, Angle,
30x 15ml RB/FA,
ID
Blank
NO Bucket/Adapter
12.5
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.3
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.5
-2.0
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
9.0
-3.5
30314834
Rotor, Angle,
12x 15ml RB/FA,
ID
Blank
NO Bucket/Adapter
11.0
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.6
-0.4
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
9.1
-1.9
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
7.7
-3.4
30314835
Rotor, Angle,
44x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.7
-0.7
30314836
Rotor, Angle,
30x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
9.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
9.1
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
8.4
-1.0
30314838
Rotor, Angle,
24x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314839
Rotor, Angle,
12x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
6.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
6.3
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
5.6
-0.9
EN-46 Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314900
Adapter, 1x 1.5ml for ov er16000x g,
6/pk
6.4
-0.1
30553085
Rotor Sw ingOut
4x 750ml D99mm
ID SL
Blank
NO Bucket/Adapter
19.3
30553138
Adapter 32x 7ml D13mm RBx 2 V1
18.7
-0.6
30553135
Adapter 18x 15ml D17mm RBx 2 V1
18.8
-0.5
30553132
Adapter 8x 30ml D26mm RBx 2 V1
19.1
-0.2
30553131
Adapter 7x 50ml D29mm RBx 2 V1
19.1
-0.2
30553127
Adapter 3x 100ml D41mm RBx 2 V1
19.1
-0.2
30553125
Adapter 1x 250ml D62mm FBx 2 V1
18.9
-0.4
30553126
Adapter 1x 225/190ml D61mm FAx 2
V1
19.1
-0.2
30553128
Adapter 5x 50ml D29mm FAx 2 V1
19.1
-0.2
30553130
Adapter 7x 50ml D29mm FAx 2 V1
19.1
-0.2
30553133
Adapter 14x 15ml D17mm FAx 2 V1
19.2
-0.1
30553134
Adapter 17x 15ml D17mm FAx 2 V1
19.0
-0.3
30553140
Adapter 18x 5/10ml D16mm
Vac/Sarx 2 V1
19.0
-0.3
30559377
Adapter 38x 1.1/1.4ml D8mm Sarx2
V1
18.9
-0.4
30553136
Adapter 21x 10ml D16mm RBx 2 V1
18.9
-0.4
30553139
Adapter 27x 1.8/7ml D13mm
Vac/Sarx 2 V1
18.8
-0.5
30553086
Rotor Angle
4x 500ml D69mm
ID
Blank
NO Bucket/Adapter
14.5
30564850
Adapter 1x 250ml D61.5mm FBx 2
V1
13.7
-0.8
30559417
Adapter 1x 175/225ml D61.5mm
FAx 2 V1
13.4
-1.1
30559419
Adapter 1x 50ml D29.5mm FAx 2 V1
12.4
-2.1
30559416
Adapter 1x 250ml D60.5mm CBCx 2
V1
12.6
-1.9
30559421
Adapter 4x 30ml D25.5mm RBx 2 V1
13.8
-0.7
30559420
Adapter 3x 50ml D29mm RBx 2 V1
14.3
-0.2
30559422
Adapter 6x 15ml D17mm FAx 2 V1
13.8
-0.7
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-47
11.7 Table 8: Redemption form / Decontamination certificate
Enclose this form with all returns of equipment and assemblies!
The completed declaration about the decontamination is a prerequisite for the assumption and further processing of the return. If no
corresponding explanation is enclosed, we carry out decontamination with costs at your expense.
Surname; last name: ___________________________________________
Organization / company: ________________________________________
Street: ______________________________________________________
ZIP CODE: _____________ place:____________________
Telephone: _____________ fax:____________________
E-Mail: _____________________________________
Pos. Crowd
Decontaminated
object
Serial number
Description /
Comment
1
2
3
4
Are the parts listed above in contact with the following substances?
Health endangering watery solutions, buffers, acids, alkalis:……………………..... Yes No
Potentially infectious agents: …………………………………………………………… Yes No
Organic reagents and solvent: ……………………………………………………….. .. Yes No
Radioactive substances: ……………………………………………. α.. β.. γ Yes No
Health endangering proteins: ………………………………………………………….. Yes No
DNA: ……………………………………………………………………………………… Yes No
These substances have reached the equipment/assembly? ………………………. Yes No
Which one, if yes:
________________________________________________________________
Description of the measures for the decontamination of the listed parts:
I confirm the proper decontamination:
Company/Dept .________________________ Place and Date: ____________________________________
Signature of the authorized person:________________________________________
Please fill out in block
capitals!
Croquis del FC5714 / FC5718 / FC5816 / FC5916
Figura 1
Croquis del FC5718R
Figura 2
Croquis del FC5816R / FC5916R
Figura 3
1 Tapa de la centrífuga
2 Ventana del rotor
3 Pantalla
4 Etiqueta de funciones
5 Interruptor de corriente principal
6 Conexión de enera
7 Liberación de emergencia
6
1
2
4
7
3
4
1
5
6
7
2
3
4
1
5
6
7
2
3
Etiqueta de funciones
Etiqueta de funciones para FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figura 4
Figura 5
1. Pantalla LCD
2. Configuración de programa
3. Inicio de centrifugación
4. Selección de RPM/RCF
5. Selección de intensidad
Aceleración/Desaceleración
6. Configuración de temperatura
(Solo FC5515R)
7. Configuración de tiempo
8. Guardar información de
configuración.
9. Centrifugación rápida
10. Abrir puerta
11. Detener centrifugación /
configuración
12. Perilla de ajuste
Pantalla LCD
La siguiente imagen muestra los elementos individuales de la pantalla LCD.
Figura 6
1
2
3
4
5
6
7
8
9
10
11
12
Figura 7
Campos de la pantalla:
A-1 Campo de pantalla "RPM/RCF"
A-2 Campo de pantalla "Acc/Dec" "Service"
A-3 Campo de pantalla "Time/Prog"
A-4 Campo de pantalla –"Temp"
Mensajes/logotipos de los campos de la pantalla
M1 "Cerrado"
M2 "Abierto"
M3 "Rotor"
M4 "Número de
rotor"
M5 "Revoluciones por
Minuto"
M6 "Fuerza de
centrifugación
relativa"
M7 "Aceleración"
M8 "Desaceleración"
M9 "Radio"
M10 "Programa"
M11 "Error"
M12 "Servicio"
M13 "h m s"
M14 "temperatura"
M15 "preenfriamiento"
M14
M5
M6
M4
A-1
A-2
A-3
A-4
M2
M3
M1
M7
M8
M12
M15
M9
M11
M10
M13
Tabla de meros de rotor
Indicador
del n.° del
rotor
N.° de
orden
Capacidad
Compatibilidad
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
61
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
20
30314820
Rotor, Swing out, 4x250ml, ID
21
30314821
Rotor, Angle, 6x250ml FB, ID
22
30314822
Rotor, Swing out, 4x100ml, ID
23
30314823
Rotor, Swing out, 4x100ml, ID, Sealable
24
30314824
Rotor, Swing out, 2x3MTP, ID
25
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
26
30314826
Rotor, Angle, 6x85ml RB, ID
27
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
28
30314828
Rotor, Swing out, 16x50ml, ID
29
30314829
Rotor, Angle, 10x50ml FA, ID
30
30314830
Rotor, Angle, 6x50ml RB/FA, ID
31
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
32
30314832
Rotor, Angle, 30x15ml RB/FA, ID
33
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
34
30314834
Rotor, Angle, 12x15ml RB/FA, ID
35
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
36
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
38
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
39
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
40
30314840
Rotor, Angle, 64x0.5ml, ID
41
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
85
30553085
Rotor SwingOut 4x750ml D99mm ID SL
86
30553086
Rotor Angle 4x500ml D69mm ID
TABLA DE CONTENIDO
1. INTRODUCCIÓ N ........................................................................................................................................................................... 1
1.1 DESCRIPCIÓ N ................................................................................................................................................................................... 1
1.2 CARACTERÍSTICAS ............................................................................................................................................................................. 1
1.3 DEFINICIÓ N DE LETREROS Y SÍMBOLOS DE ADVERTENCIA......................................................................................................................... 1
1.4 PRECAUCIONES DE SEGURIDAD ........................................................................................................................................................... 2
1.4.1 Usuario ................................................................................................................................................................................ 2
1.4.2 Rotor y accesorios .............................................................................................................................................................. 2
1.4.3 Medidas para su protección.............................................................................................................................................. 2
1.4.4 Excluya las siguientes influencias ambientales .............................................................................................................. 3
1.4.5 Medidas de seguridad operativa...................................................................................................................................... 3
1.4.6 Peligro y precauciones....................................................................................................................................................... 3
2. INSTALACIÓN ............................................................................................................................................................................... 4
2.1 DESEMPACADO ................................................................................................................................................................................ 4
2.1.1 Paquete de entrega ........................................................................................................................................................... 4
2.2 SELECCIÓ N DE LA UBICACIÓ N .............................................................................................................................................................. 4
2.3 INSTALACIÓ N ................................................................................................................................................................................... 4
2.4 PRECAUCIONES DE SEGURIDAD DURANTE LA OPERACIÓ N Y LA GARANTÍA .................................................................................................. 5
3. OPERACIÓN .................................................................................................................................................................................. 5
3.1 MONTAJE Y CARGA DEL ROTOR ........................................................................................................................................................... 5
3.1.1 Instalación de rotores ........................................................................................................................................................ 5
3.1.2 Carga de los rotores angulares......................................................................................................................................... 6
3.1.2 Carga de los rotores oscilantes......................................................................................................................................... 6
3.1.4 Carga y sobrecarga de rotores.......................................................................................................................................... 7
3.1.5 Remoción del rotor ............................................................................................................................................................ 7
3.2 CONTROL DE LA TAPA ........................................................................................................................................................................ 7
3.2.1 Tapa abierta ....................................................................................................................................................................... 7
3.2.2 Seguro de la tapa ............................................................................................................................................................... 8
3.3 PRESELECCIÓ N ................................................................................................................................................................................. 8
3.3.1 Preselección de velocidad / valor RCF.............................................................................................................................. 8
3.3.2 Preselección del tiempo de operación ............................................................................................................................. 9
3.3.3 Preselección de intensidad del freno y aceleración........................................................................................................ 9
3.3.4 Preselección de temperatura (solo para modelos refrigerados)................................................................................. 10
3.3.5 Preenfriamiento (solo para modelos refrigerados) ...................................................................................................... 10
3.4 CORRECCIÓ N DEL RADIO .................................................................................................................................................................. 10
3.5 PROGRAMA ................................................................................................................................................................................... 11
3.5.1 Almacenamiento de programas ..................................................................................................................................... 11
3.5.2 Recuperación de programas almacenados ................................................................................................................... 12
3.5.3 Salir del modo de programa ........................................................................................................................................... 12
3.6 ARRANQUE Y PARO DE LA CENTRÍFUGA .............................................................................................................................................. 12
3.6.1 Arranque de la centrífuga ............................................................................................................................................... 12
3.6.2 Botón "STOP".................................................................................................................................................................... 13
3.7 DETECCIÓ N DE DESBALANCE............................................................................................................................................................. 13
4. CONFIGURACIÓN ....................................................................................................................................................................... 13
4.1 AJUSTES BÁSICOS............................................................................................................................................................................ 13
4.1.1 Acceso al modo "Operating Data" ................................................................................................................................. 13
4.1.2 Indicacn de temperatura.............................................................................................................................................. 14
4.1.3 Encendido y apagado de la señal ................................................................................................................................... 15
4.1.4 Preselección del volumen de la señal de sonido ........................................................................................................... 15
4.1.5 Selección de meloa para la señal del sonido, final de operación ............................................................................ 15
4.1.6 Encendido y apagado del sonido del teclado ................................................................................................................ 16
4.1.7 Extraccn de datos operativos ...................................................................................................................................... 16
5. MANTENIMIENTO...................................................................................................................................................................... 17
5.1 MANTENIMIENTO Y LIMPIEZA ........................................................................................................................................................... 17
5.1.1 Cuidado general: ............................................................................................................................................................. 17
5.1.2 Limpieza y desinfección de la unidad............................................................................................................................. 18
5.1.3 Limpieza y desinfección del rotor ................................................................................................................................... 18
5.1.4 Desinfección de rotores de aluminio.............................................................................................................................. 18
5.1.5 Desinfección de rotores de PP......................................................................................................................................... 18
5.1.6 Ruptura de vidrio ............................................................................................................................................................. 19
5.2 VIDA ÚTIL DE ROTORES, PORTATUBOS Y ACCESORIOS ........................................................................................................................... 19
6. IDENTIFICACIÓN Y SOLUCIÓ N DE PROBLEMAS ...................................................................................................................... 19
6.1 MENSAJE DE ERROR: CAUSA / SOLUCIÓ N........................................................................................................................................... 19
6.2 ESTUDIO DE POSIBLES MENSAJES DE ERROR Y SUS SOLUCIONES.............................................................................................................. 19
6.2.1 Apertura de la tapa durante una falla de la corriente eléctrica (Apertura de emergencia de la tapa) ................. 19
6.2.2 Descripción del sistema de mensajes de error .............................................................................................................. 20
7. RECEPCIÓN DE CENTRÍFUGAS PARA REPARACIÓ N ................................................................................................................ 21
8. TRANSPORTE, ALMACENAMIENTO Y ELIMINACIÓN.............................................................................................................. 21
8.1 TRANSPORTE.................................................................................................................................................................................. 21
8.2 ALMACENAMIENTO......................................................................................................................................................................... 21
9. FICHA TÉCNICA........................................................................................................................................................................... 23
9.1 ESPECIFICACIONES .......................................................................................................................................................................... 23
9.1.1 Centrifuge FC5714 ............................................................................................................................................................ 23
9.1.2 Centrífuga FC5718............................................................................................................................................................ 24
9.1.3 Centrifuge FC5718R.......................................................................................................................................................... 25
9.1.6 Centrifuge FC5916 ............................................................................................................................................................ 28
9.1.7 Centrifuge FC5916R.......................................................................................................................................................... 29
9.2 DIAGRAMAS Y DIMENSIONES ................................................................................................................................................... 30
10. CONFORMIDAD..................................................................................................................................................................... 31
11. APÉNDICE............................................................................................................................................................................... 32
11.1 TABLE 1: PESO NETO PERMITIDO ............................................................................................................................................. 32
11.2 TABLE 2: TEMPERATURAS MÍNIMAS A MÁXIMA VELOCIDAD ͘͘͘͘.........................................................................................͘͘34
11.3 TABLE 3: MÁXIMA VELOCIDAD Y VALORES RCF PARA ROTORES PERMITIDOS............................................................͘͘͘͘͘͘͘͘͘35
11.4 TABLE 4: TIEMPOS DE ACELERACIÓ N Y DESACELERACIÓ N ..................................................................................................... 38
11.5 TABLA 5: MENSAJES DE ERROR....................................................................................................................................................... 40
11.6 TABLE 6: CORRECC N DEL RADIO ........................................................................................................................................... 41
11.7 TABLA 7: FORMULARIO DE DEVOLUCIÓ N / CERTIFICADO DE DESCONTAMINACIÓ N ................................................................................. 45
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-1
1. INTRODUCCIÓN
1.1 Descripción
Gracias por elegir este producto OHAUS.
Todos los símbolos indican instrucciones de seguridad y puntos de situaciones peligrosas potenciales.
Lea el manual por completo antes de usar las centfugas Frontier
TM
MultiPro para evitar un funcionamiento
incorrecto.
Las centrífugas Frontier
TM
MultiPro fueron disadas para la separación de materiales o mezclas con diferentes
densidades.
Las centrífugas OHAUS están diseñadas exclusivamente para funcionar en interiores y para ser utilizadas solo por
personal cualificado.
1.2 Caractesticas
El modelo FC5714/FC5718/FC5816/FC5916 es una centfuga universal no refrigerada.
El modelo FC5718R/FC5816R/FC5916R es una centrífuga universal refrigerada.
Los tres modelos que ofrecemos pueden funcionar en dos voltajes, 230 V o 120 V.
Las centfugas pueden utilizarse con rotores angulares y basculantes.
Todos los parámetros son accesibles a través de botones y se seleccionan con el ajustador central. Todos los
valores preseleccionados y actuales se muestran permanentemente en la pantalla LCD.
La centrífuga es impulsada por un motor de inducción libre de mantenimiento.
Los datos técnicos detallados se pueden encontrar en la sección "Ficha técnica".
1.3 Definición de letreros y símbolos de advertencia
Las notas de seguridad están marcadas con palabras de señalamiento y símbolos de advertencia. Estos muestran
problemas y advertencias de seguridad. Ignorar las notas de seguridad puede resultar en lesiones, daño al
instrumento, mal funcionamiento y resultados falsos.
El grado de peligro es parte de una nota de seguridad y distingue los posibles resultados de la falta de
observancia.
Palabras de señalamiento
PELIGRO
Conducirá a lesiones graves o muerte si no se evita.
ADVERTENCIA
Para una situación peligrosa con riesgo medio, que posiblemente resulta en
lesiones o muerte si no se evita.
PRECAUCIÓN
Para una situación peligrosa con riesgo bajo, que resulta en daño al
dispositivo,
a la propiedad o en rdida de datos si no se evita.
ATENCIÓN
Para información importante acerca del producto. Puede resultar en daño del equipo
si no se evita.
NOTA
Para información útil acerca del producto.
mbolos de advertencia
Riesgo general Riesgo de descarga eléctrica
Corriente alterna Riesgo biológico
Explosión Aplastamiento
ES-2 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Letreros de advertencia e información en la superficie de la centrífuga
Se deben usar cuatro portatubos siempre en los rotores basculantes de cuatro
lugares, o de lo contrario ocurridaño a la centfuga. Este daño no esta
cubierto por la garantía del producto.
¡Atención! Revise el apriete de la tuerca del rotor antes de cada uso.
Desconecte el enchufe de la corriente principal antes de abrir la carcasa o la
apertura de emergencia.
Direccn de rotación el impulsor del rotor tiene rotación a la derecha
Referencia para carga de rotores
1.4 Precauciones de seguridad
1.4.1 Usuario
Las centrífugas OHAUS están diseñadas exclusivamente para uso en interiores y para uso por parte de personal
calificado. Este dispositivo solo puede ser usado por personal especialista capacitado. Dicho personal debe haber
leído el manual de operación y estar familiarizado con el funcionamiento del dispositivo.
1.4.2 Rotor y accesorios
Solo deben usarse rotores y accesorios OHAUS originales. Cualquier otro uso o uso intencional se considera
inadecuado. OHAUS no se hace responsable por daños que resulten del uso inadecuado.
PRECAUCIÓN:
Lea todas las advertencias de seguridad antes de instalar, hacer conexiones o dar servicio a este
equipo. La falta de cumplimiento de estas advertencias podría resultar en lesiones o daños
a la propiedad. Conserve las instrucciones para referencia futura.
1.4.3 Medidas para su protección
ADVERTENCIA: ¡Nunca trabaje en un entorno sujeto a riesgos de explosión! La caja del instrumento
no es hermética al gas. (Riesgo de explosn debido a la generación de chispas y corrosión causada
por el ingreso de gases.)
ADVERTENCIA: Cuando use químicos y solventes, siga las instrucciones del fabricante y las reglas
generales de seguridad de laboratorio.
ADVERTENCIA: La centrífuga no está sellada. Adopte medidas de protección adecuadas cuando use
la centrífuga para muestras infecciosas o patógenas. Tome precauciones de seguridad adecuadas
cuando maneje muestras de estos tipos.
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-3
1.4.4 Excluya las siguientes influencias ambientales
Vibraciones fuertes
Luz solar directa
Humedad atmosférica mayor del 80%
Presencia de gases corrosivos
Temperaturas menores de 2 °C y mayores de 35 °C
Campos eléctricos o magticos intensos:
ADVERTENCIA:
Existe el riesgo de descarga eléctrica dentro de la caja. Solo personal autorizado y calificado debe
abrir la caja. Desconecte todas las conexiones eléctricas hacia la unidad antes de abrirla.
1.4.5 Medidas de seguridad operativa
No destornille las dos mitades de la caja.
¡Seque cualquier derrame de quidos inmediatamente! El instrumento no es hermético al agua.
Verifique que el rango de voltaje de entrada y tipo de enchufe sean compatibles con la alimentacn eléctrica
local.
Conecte el cable de corriente únicamente a un enchufe adecuado con toma de tierra.
Utilice únicamente un cable de corriente con una potencia que supere las especificaciones del etiquetado del
equipo.
No coloque el equipo de forma que resulte difícil desconectar el cable de corriente del enchufe.
Verifique que el cordón eléctrico no represente un obstáculo potencial o riesgo de tropezarse.
El equipo es solo para uso en interiores. Use el equipo solamente en lugares secos.
Use solo accesorios aprobados.
Opere el equipo sólo en las condiciones ambientales especificadas en estas instrucciones
Desconecte el equipo de la corriente eléctrica cuando lo limpie.
No opere el equipo en ambientes peligrosos o inestables.
El servicio debe proporcionarse solamente por personal autorizado.
1.4.6 Peligro y precauciones
Para proteger a las personas y el entorno, siga estas precauciones:
Durante la centrifugación, se probe la presencia de personas y la manipulación de materiales peligrosos
dentro de 30 cm alrededor de la centfuga, de acuerdo con las regulaciones de EN 61010-2-020.
La FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R no es a prueba de explosn y por lo tanto no
debe operarse en áreas o lugares con riesgo de explosión. La centrifugacn de sustancias inflamables,
explosivas, radioactivas o similares, que reaccionan en forma qmica con la enera, está estrictamente
prohibida. La decisión final sobre los riesgos asociados con el uso de tales sustancias es responsabilidad del
usuario de la centfuga.
Nunca centrifugue material tóxico o patógeno sin las precauciones de seguridad adecuadas; por ejemplo, la
centrifugación de portatubos o tubos cuyo sello hermético no está presente o es defectuoso está estrictamente
prohibida. El usuario está obligado a implementar procedimientos de desinfección adecuados en caso de que
sustancias peligrosas contaminen la centrífuga o sus accesorios. Cuando centrifugue sustancias infecciosas,
siempre preste atencn a las precauciones generales de laboratorio. ¡Si es necesario, ngase en contacto
con su representante de seguridad!
Se prohíbe operar la centrífuga con rotores que no sean los indicados para esta unidad.
Bajo ninguna circunstancia abra la tapa de la centfuga mientras el rotor esté n funcionando o girando a una
velocidad > 2 m/s.
ES-4 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
1.4.7 Abreviaturas usadas en este manual
Símbolo/Abrev iaturas
Unidad
Descripción
RPM
[min
-1
] rpm
revoluciones por minuto
RCF
[x g]
fuerza centrífuga relativa
PCR
PCR Reacción en cadena de la polimerasa
PP
-
Polipropileno
PC
-
Policarbonato
accel
-
aceleración
decel
-
desaceleración
Prog
-
programa
2. INSTALACIÓN
2.1 Desempacado
Retire con cuidado la centfuga y cada uno de sus componentes del paquete. Los componentes incluidos varían
de acuerdo con el modelo de centrífuga (vea la tabla siguiente). Guarde el empaque para poder almacenar y
transportar la centrífuga adecuadamente. ¡El manual de instrucciones debe mantenerse siempre con la centfuga!
Los rotores y accesorios se empaquetan aparte.
ADVERTENCIA: Peligro de levantamiento. Existe riesgo de lesiones si una sola persona la
levanta. Solicite ayuda al levantar o mover el equipo.
Consulte la seccn 8.3 para obtener detalles sobre cómo sacar el equipo del embalaje.
Se recomienda que cuatro o más personas transporten e instalen estas tres unidades, o utilicen aln
equipo de transporte. Consulte la sección 8.3 para ver más detalles sobre cómo levantar estas unidades
cuando esn con embalaje.
2.1.1 Paquete de entrega
2.2 Selección de la ubicacn
¡NOTA!
Evite las vibraciones excesivas, fuentes de calor, corrientes de aire y cambios pidos de temperatura.
La centrífuga debe instalarse en una superficie sólida, nivelada y uniforme, si es posible en un
gabinete o mesa de laboratorio, o en alguna otra superficie libre de vibraciones.
Durante la centrifugación, la centrífuga debe colocarse de tal forma que quede un espacio mínimo de 30 cm en
cada lado de la unidad, de acuerdo con los estándares EN 61010-2-020.
No coloque la centrífuga cerca de una ventana o calentador donde pudiera estar expuesta al calor excesivo, ya
que el funcionamiento de esta se basa en una temperatura ambiente de 23 °C.
2.3 Instalación
Siga estos pasos:
Verifique si la corriente eléctrica corresponde a la especificada en la etiqueta de clasificación del fabricante, la
cual está en el panel posterior.
Para el modelo FC5714/FC5718/FC5816, la nea de alimentación debe estar protegida por un disyuntor de 10
A circuito de calificación (tipo K).
Para los modelos FC5718R/FC5816R /FC5916/FC5916R, la línea de alimentación debe estar protegida por un
disyuntor Una calificación de 16 (tipo K).
Para caso de emergencia, debe haber un interruptor de emergencia instalado fuera de la sala para desconectar
la corriente eléctrica de la unidad.
Conecte la centfuga en la toma de corriente eléctrica con conexión a tierra.
Conecte la centrífuga al suministro eléctrico.
Cantidad
Descripción
1
Centrífuga
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
1
Cordón eléctrico
1
Tarjeta de garantía
1
Manual de instrucciones / Guía rápida
1
Llave del rotor
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-5
(La toma para el cable de alimentación debe ser fácil de alcanzar y también fácil de desconectar)
Encienda el instrumento usando el interruptor de corriente eléctrica.
Abra la tapa con el botón "Door Open".
Retire el dispositivo de fijación de transporte del motor.
2.4 Precauciones de seguridad durante la operación y la garantía
No opere la centrífuga en caso de que no esté instalada correctamente.
Durante su funcionamiento, no se apoye en la centrífuga.
Por razones operativas, no permanezca más de lo necesario en el radio de 30 cm alrededor del equipo.
No coloque ningún material potencialmente peligroso en el radio de 30 cm alrededor del equipo.
No opere la centrífuga cuando esté desarmada (por ejemplo, sin la caja).
No opere la centrífuga cuando se hayan alterado ilegalmente componentes mecánicos o eléctricos.
No use accesorios como rotores y portatubos que no estén aprobados exclusivamente por OHAUS Corporation,
excepto tubos para centfuga disponibles comercialmente hechos de vidrio o plástico.
No centrifugue sustancias sumamente corrosivas, ya que pueden dañar o debilitar los materiales.
No opere la centrífuga con rotores o portatubos que muestren signos de corrosn o daño mecánico.
El fabricante es responsable de la seguridad y confiabilidad de la centrífuga solamente si:
La unidad se opera de acuerdo con este manual de instrucciones.
Las modificaciones, reparaciones u otros ajustes son realizados por personal autorizado y la instalación
eléctrica cumple con el código eléctrico correspondiente.
ATENCIÓN:
Garantía
La centrífuga ha sido sometida a rigurosas pruebas y controles de calidad. En el improbable caso de que
ocurra una falla por defecto de fabricación, la centrífuga y los rotores están cubiertos por una garantía. Esta
garantía queda sin validez en caso de mal manejo, dos o negligencia, y además en el caso de utilizar piezas
de repuesto y/o accesorios incorrectos o realizar modificaciones no autorizadas en la unidad.
El fabricante se reserva el derecho de realizar modificaciones técnicas para mejorar el funcionamiento de la
unidad.
3. OPERACIÓN
3.1 Montaje y carga del rotor
3.1.1 Instalación de rotores
Limpie el eje motriz y el collar con una tela limpia y libre de grasa. Coloque el rotor en el eje motriz. (Vea la figura
siguiente). Tenga cuidado de que el rotor esté completamente instalado en el eje del motor.
Figura 8
Eje del motor y cámara
ES-6 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figura 9 Figura 10
Sostenga el rotor con una mano y fíjelo en el eje al girar la tuerca de fijación en sentido del reloj. Apriete la tuerca
de fijación con la llave del rotor incluida (vea las figuras 9-10)
Proporcionaremos una herramienta para rotor sin tuerca con la centrífuga; la herramienta para rotor con tuerca se
proporcionará con el rotor.
ATENCIÓN:
Verifique que el tornillo de fijación esté instalado correctamente antes de cada operación. (Vea la
figura 11.)
No opere la centrífuga con rotores o portatubos que muestren
signos de corrosión o do mecánico.
No la opere con sustancias sumamente corrosivas que
pudieran dañar el rotor, los portatubos y los materiales.
En caso de alguna duda, ngase en contacto con el
fabricante.
3.1.2 Carga de los rotores angulares
Los rotores deben cargarse en forma simétrica y con pesos iguales (vea la figura siguiente). El adaptador solo
puede cargarse con los portatubos adecuados. Las diferencias en peso entre los portatubos dañados deben
mantenerse lo más pequeñas que sea posible. Por lo tanto, se recomienda pesarlos con una balanza. Esto reduce
el desgaste del motor y el ruido de operación.
La carga máxima por orificio está establecida para cada rotor.
Figura 11 INCORRECTO Figura 12 CORRECTO (6 tubos)
3.1.2 Carga de los rotores oscilantes
La carga de los portatubos debe hacerse de acuerdo a la figura a continuación.
Espermitido operar por ejemplo un rotor de 4 lugares con 2 portatubos cargados solamente, pero los portatubos
cargados deben estar opuestos entre ellos. Asegúrese de que los portatubos no cargados también se coloquen
dentro del rotor (ver la figura abajo).
En principio, los rotores basculantes no se pueden hacer funcionar hasta que todos los portatubos o racks se
coloquen en el rotor.
Los pernos en el rotor deben lubricarse con aceite "High TEF". Los tubos de muestra deben llenarse todos hasta el
mismo nivel y deben colocarse en los huecos o racks de tubos. La diferencia de peso de los portatubos cargados
no debe exceder aproximadamente 1.0 g.
Tuerca para rotor
Herramienta
para rotor
con tuerca
Herramienta
para rotor sin
tuerca
Tapa
insertable
Tapa
atornillable
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-7
ATENCIÓN
Los rotores basculantes solo pueden hacerse funcionar si todas las posiciones están llenas con
cuatro portatubos o cuatro portavasos ¡¡no mezcle portatubos y portavasos!!
ATENCIÓN
No opere la centfuga con rotores o portatubos que muestren cualquier signo de corrosión o do
mecánico.
No opere la unidad con sustancias extremadamente corrosivas, que podrían dañar el rotor y los
portatubos.
En caso de cualquier pregunta, póngase en contacto con el fabricante.
Figure.13 WRONG Figure.14 CORRECT
3.1.4 Carga y sobrecarga de rotores
Todos los rotores aprobados con su máxima velocidad y peso de llenado máximo están indicados en la "Tabla 2,
peso neto permitido" (vea el APÉNDICE).
La carga máxima permitida para un rotor, determinada por el fabricante, así como la máxima velocidad permitida
para el rotor (vea la etiqueta en el rotor), no deben excederse. Los quidos con los que se cargan los rotores
deben tener una densidad homogénea máxima de 1.2 g/ml o menos cuando el rotor opera a máxima velocidad.
Para centrifugar quidos de mayor densidad, la velocidad debe reducirse de acuerdo con la siguiente fórmula:
Vel. reducida n
red
=
1,2
𝑑𝑒𝑛𝑠𝑖𝑑𝑎𝑑 𝑚𝑎𝑦𝑜𝑟
x máx. vel. (n
máx
) del rotor
Ejemplo:
n
red
=
1,2
1,7
x 4.000 = 3.360 rpm
En caso de alguna duda, póngase en contacto con el fabricante.
Para determinar la fuerza centrífuga relativa (RCF/fuerza G) para un adaptador específico, puede calcular
aplicando DIN 58 970 usando la fórmula siguiente:
RCF = 1.117862 *10
5
*n
2
*r
max
n: revoluciones por minuto (RPM)
r
max
: radio de centrifugado máximo en cm usando la parte inferior de los tubos
3.1.5 Remoción del rotor
Afloje completamente la tuerca de fijación (tornillo sobre el punto gido) y levante el rotor en forma vertical fuera
de la centrífuga. (Vea las figuras 9 y 10).
3.2 Control de la tapa
3.2.1 Tapa abierta
Después de la operación, cuando la tapa de la centrífuga está cerrada, aparece la palabra "close" (M1) en la
pantalla "RPM | RCF" (A-1). Además, si hay un rotor en la centrífuga, aparece la palabra "rotor"(M3) así como el
mero de código del rotor respectivo, el cual está en el sistema de la centrífuga "71" (M4). Si no hay rotor en la
centfuga, destella la palabra "rotor" (M3) y aparece adicionalmente la palabra "no" (M4). Al presionar el botón
"Door Open" (7), puede liberar la tapa de la centfuga. Tan pronto la tapa electromagnética esté completamente
liberada, aparece la palabra "open" (M2). Ahora puede abrir la tapa de la centrífuga. Consulte la Figura 15 a
continuación para referencia.
ES-8 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figura 15
Durante la operación, puede consultar el tipo de rotor en cualquier momento al presionar el botón "Door Open"
(7).
3.2.2 Seguro de la tapa
La tapa debe bajarse solo un poco. Un seguro de la tapa electromagnética cierra la tapa, al mismo tiempo que
desaparece la palabra "open" (M2) (vea la Figura 15). Como signo de que la centrífuga está lista para arrancar,
en la pantalla "RPM | RCF" (A-1) aparece la palabra "close" (M1). Al mismo tiempo, aparece la palabra "rotor"
(M3), así como el número de código del rotor, el cual está en el sistema de la centfuga, "no 71" (M4). Con eso,
todos los datos específicos del rotor, por ejemplo máxima velocidad, aceleración, etc. están adoptados.
ATENCIÓN: ¡No deje los dedos entre la tapa y el dispositivo o mecanismo de cierre cuando cierre la
tapa!
3.3 Preselección
3.3.1 Preselección de velocidad / valor RCF
Esta preselección se activa con el botón "RPM | RCF" (4) (consulte la Figura 16 a continuación). Al presionar el
botón una vez, destella la palabra "rpm" (M5). Al presionar el botón dos veces, se puede seleccionar la
preselección de las fuerzas centrífugas. Entonces aparece la palabra "rcf" (M6). Puede establecer los valores que
desee con la perilla de ajuste (1). En la pantalla (A-1) aparece permanentemente el valor regulado, antes, durante
y después de la operación.
Figura 16
Mientras no haya rotor insertado, la velocidad es ajustable entre 200 rpm y la máxima revolución de la centrífuga.
Si hay un rotor en la centrífuga, la velocidad solo puede preseleccionarse hasta la máxima revolución permitida de
ese rotor. Esta es la misma que la preselección del valor RCF. El rango de ajuste es entre la fuerza centrífuga a
200 rpm y la fuerza centrífuga máxima permitida del rotor.
Vea la "Tabla 4: máxima velocidad y valores RCF para rotor permitido" (Vea el APÉNDICE). Todos los
valores importantes se indican aquí.
7
1
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-9
ATENCIÓN:
También verifique las revoluciones máximas permitidas de sus tubos de prueba con el fabricante.
3.3.2 Preselección del tiempo de operación
El tiempo de operación puede preseleccionarse en tres rangos diferentes desde 10 segundos hasta 99 horas 59
minutos.
1. Rango desde 10 segundos hasta 59 minutos 50 segundos en pasos de 10 segundos
2. Rango desde 1 hora hasta 99 horas 59 minutos en pasos de 1 minuto
3. La operación continua "cont", que puede interrumpirse con el botón "Stop"(10) (vea la Figura 17).
El tiempo de operación puede preseleccionarse con la tapa abierta o cerrada.
Para activar la configuración del tiempo de operación, presione el botón "Time" (6).
En la pantalla "Time/Prog" (A-3) destella la indicación "m : s" o "h : m", dependiendo de la configuración previa.
Para establecer el valor deseado, use la perilla de ajuste (1). Después de exceder 59 min. 50 seg., la indicación
cambia automáticamente a "h : m". Después de exceder 99 horas 59 min., aparece la palabra "cont" en la
pantalla "Time/Prog" (A-3). Esta operación continua solo puede interrumpirse al presionar el botón "Stop" (10). El
conteo regresivo del tiempo inicia una vez que se alcanza la velocidad establecida.
La pantalla siempre muestra el tiempo de operación restante. (Vea la figura 17.)
Figura 17
3.3.3 Preselección de intensidad del freno y aceleración
Esta función se activa con el botón "Accel/Decel" (5) (vea la Figura 18).
Al presionar el botón una vez, destella la palabra "accel" (M7) en la pantalla "Acc/Dec" (A-2). La aceleración
deseada puede preseleccionarse con la perilla de ajuste (1). El valor 0 equivale a la menor y el valor 9 a la mayor
aceleración.
Al presionar dos veces el botón "Accel/Decel" (5), la pantalla "Acc/Dec" (A-2) muestra la palabra "decel"(M8).
Ahora puede preseleccionarse la intensidad del freno con la perilla de ajuste (1). El valor 9 equivale al más corto y
el valor 0 al tiempo de freno más largo posible.
Vea la "tabla 5: tiempos de aceleración y desaceleración" (APÉNDICE). En esta se muestran los tiempos de
aceleración y desaceleración para las etapas de aceleración y desaceleración 0 a 9 para rotores permitidos.
Figura 18
ES-10 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
3.3.4 Preselección de temperatura (solo para modelos refrigerados)
Esta función se activa con el botón "Temp/Setup" (13). Después de presionar este botón, en la pantalla
"Time/Prog" destella la indicación "°C" (A-4). Con la perilla (1) puede preseleccionarse la temperatura de prueba
deseada en pasos de 1 °C en un rango de -20 °C hasta +40 °C.
El valor se indica permanentemente en la pantalla (Figura 19) antes, durante y después de la operación.
¡Observe las mínimas temperaturas respectivas de los rotores a máxima velocidad
Figura 19
3.3.5 Preenfriamiento (solo para modelos refrigerados)
Si las muestras son sensibles a la temperatura, es útil preenfriar la centrífuga, el rotor y eventualmente los
portatubos a la temperatura de trabajo requerida. Por lo tanto, inserte el rotor deseado y preajuste la temperatura
respectiva. Al presionar en forma simultánea los botones "Temp/Setup" (13) (vea la Figura 20) y "Time" (6), inicia
la operación. Durante la operación, la unidad selecciona automáticamente una velocidad rotacional que es
equivalente al 20% de la velocidad rotacional permitida del rotor respectivo. Después de que se alcanza la
temperatura predeterminada, puede salir de la operación de preenfriamiento con el botón "Stop" (10).
Dependiendo del rotor insertado, el preenfriamiento dura aproximadamente entre 10 y 20 minutos.
Figura 20
3.4 Corrección del radio
Si usa adaptadores o reductores, puede cambiar el radio centrífugo del rotor respectivo. En ese caso, puede
corregir el radio manualmente. Proceda de la siguiente manera:
Cierre la tapa, luego presione el botón "Time" (6) (vea la Figura 21) y el botón "Prog/Setup" (11) al mismo tiempo
y manténgalos presionados.
En la pantalla "Time/Prog" (A-3) aparece la palabra "radius" (M9). Con la perilla de ajuste (1) puede
preseleccionar la corrección del radio respectiva (vea la Tabla 7, APÉNDICE) en pasos de 0.1 cm. Una vez que
haya establecido la corrección del radio, aparece la palabra "radius" (M9). Esta palabra será visible hasta que
vuelva a poner la corrección del radio en 0.
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-11
Figura 21
3.5 Programa
3.5.1 Almacenamiento de programas
Puede almacenar hasta 99 operaciones con todos los pametros relevantes, incluyendo los rotores usados.
Puede usar cualquier número de programa libre y volver a consultarlo.
Coloque el rotor necesario en la centfuga. Al presionar el botón "Prog/Setup" (11) en la pantalla "Time/Prog"
(A-3) aparece la palabra "programm". Con la perilla de ajuste (1) puede seleccionar el mero de programa
deseado.
Si un mero de programa ya está ocupado, en la pantalla "RPM | RCF" (A-1) aparecerán las palabras "rotor"
(M3) y "xx" (M4). En caso de meros de programa libres, aparece 0.
Figura 22
Cierre la tapa de la centrífuga. Ahora proceda como se describió anteriormente para configurar todos los
parámetros de operación importantes. Si la tapa no se cierra al almacenar el programa, en la pantalla "RPM/RCF"
(A-1) destellan las palabras "FirSt" y "CLOSE Lid" (vea la Figura 23) alternadamente. Si desea iniciar la
operación sin almacenar el programa, en la pantalla "RPM/RCF" (A-1) destellan las palabras "First" y "PrESS
StoreE" (vea la Figura 24) alternadamente.
Figura 23 Figura 24
ES-12 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Para adopción de datos, presione el botón "Store" (12) (vea las Figuras 23 y 24) durante aproximadamente 1
segundo. Si el programa se almacena correctamente, aparece la palabra "StorE" en la pantalla "RPM/RCF"
(A-1). Como resultado, desaparece la palabra "programm" (M10).
Tan pronto como se suelta el botón "Store" (12), vuelve a aparecer la palabra "programm xx" (M10); la (xx)
representa la ubicación de programa seleccionado.
Si todos los meros de programa están ocupados, puede tomar un número antiguo que ya no se necesite y
simplemente ingresar los nuevos parámetros.
3.5.2 Recuperacn de programas almacenados
Para recuperar programas almacenados, presione el botón "Prog/Setup" (11) (vea la Figura 25) mientras la tapa
está cerrada. En la pantalla "Time/Prog" (A-3) aparece "programm –"(M10). El número de programa deseado
puede preseleccionarse con la perilla de ajuste (1).
Los valores almacenados para ese programa aparecerán en las pantallas respectivas.
Si dentro de la centrífuga está el rotor equivocado para el programa seleccionado, en la pantalla "RPM | RCF" (A-
1) destella la palabra "rotor" (M3). Al mismo tiempo, la palabra "FALSE" y el número de rotor almacenado " xx"
(M4) destellan alternadamente.
Figura 25
3.5.3 Salir del modo de programa
Para salir del modo de programa, solo presione el botón "Prog/Setup" (11) (vea la Figura 25). Entonces, dentro
de la pantalla "Time/Prog" aparece la palabra "programm".
Configure la pantalla en "programm--" (M10) con la perilla de ajuste (1).
3.6 Arranque y paro de la centrífuga
3.6.1 Arranque de la centfuga
Puede arrancar la centrífuga ya sea con el botón "Start" (9) (vea la Figura 26) o con el botón "Quick Spin" (8).
Con el botón "Start" (9), puede iniciar operaciones almacenadas u operaciones con pametros preseleccionados
manualmente.
Una vez que finalice el tiempo de operación preseleccionado, la centrífuga para automáticamente.
Con el botón "Quick Spin" (8), puede iniciar operaciones que duran solo unos segundos.
Al presionar el botón "Quick Spin" (8), la centfuga se acelera hasta las revoluciones preseleccionadas.
En la pantalla "Time/Prog" (A-3), se indica el tiempo de operación transcurrido desde la fecha en que se presiona
el botón "Quick Spin" (8).
Al soltar el botón "Quick Spin" (8), la centfuga se detiene y el tiempo de operación se indica hasta la apertura de
la tapa.
Figura 26
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-13
3.6.2 Botón "STOP"
Con el botón "Stop" (10) (vea la Figura 27) puede interrumpir la operación en cualquier momento. Después de
presionar el botón, la centrífuga se desacelera con la intensidad preseleccionada en descenso hasta parar
completamente.
Figura 27
3.7 Detección de desbalance
En caso que el rotor no sea cargado de manera uniforme, el motor se apagará durante la aceleracn. El rotor se
desacelera hasta el paro total.
Cuando en la pantalla "Time/Prog" (A-3) aparezca la palabra "error” (M11) junto con el mero "01, la
diferencia de peso de las muestras es demasiado grande. Distribuya el peso de manera uniforme.
Cargue el rotor como se describió en el catulo 3.1.2 y 3.1.3.
Cuando en la pantalla "Time/Prog" (A-3) aparezca la palabra "error” junto con el mero "02" (vea la Figura 28),
puede deberse a lo siguiente: el interruptor de desbalance está defectuoso.
Figura 28
4. CONFIGURACIÓN
4.1 Ajustes sicos
4.1.1 Acceso al modo "Operating Data"
Cuando use la centrífuga, se pueden configurar los siguientes pametros:
- Indicación de temperatura en °C o °F
- Encendido/apagado de la sal acústica
- Encendido/apagado del sonido del teclado
- Preselección del volumen de la señal de sonido
- Seleccn de melodía de la señal del sonido de final de operación o "end of run"
Cuando la centrífuga está apagada, presione al mismo tiempo los botones "Time"(6) y "Door Open" (7) y
encienda el interruptor principal de la centfuga. Ahora suelte ambos botones y como resultado se ejecuta una
prueba de pantalla durante aproximadamente 5 segundos. Todos los indicadores aparecerán al mismo tiempo
(vea la Figura 29).
ES-14 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figura 29
ATENCIÓN:
Observe que debe ingresar en el programa como se describe en el punto 4.1.1 para cambiar los
ajustes de los puntos 4.1.2 a 4.1.7. Una vez que haya guardado las configuraciones,
puede cambiar al modo de programa normal nuevamente al apagar la centfuga durante un tiempo
breve.
Todas las configuraciones cambiadas deben confirmarse con el botón "Start"(9). Aparece la palabra "Store"(12)
en la pantalla "RPM | RCF"(A-1). ¡Solamente entonces son válidas las preselecciones!
Figura 30
4.1.2 Indicación de temperatura
Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón
"Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) aparece la palabra "Service". Ahora seleccione la letra "C"
con la perilla de ajuste (1). Como resultado, en la pantalla "RPM | RCF" (A-1) aparecen las palabras
"CELSI/temp". Si presiona el botón "RPM | RCF" (4), la palabra "CELSI" destella y puede cambiar la pantalla a
Fahrenheit "FAREN" con la perilla de ajuste (1) (vea la Figura 31).
Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa
normal al apagar la centrífuga durante un tiempo breve.
Figura 31
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-15
4.1.3 Encendido y apagado de la señal
Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón
"Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) destella la palabra "Service". Ahora seleccione la letra "L" con
la perilla de ajuste (1). Como resultado, aparecen las palabras "On Sound" en la pantalla "RPM | RCF" (4). Si
presiona el botón "RPM | RCF" (4) ahora, destella la palabra "On" y puede apagar el sonido con la perilla de
ajuste (1) (vea la Figura 32).
Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa
normal al apagar la centrífuga durante un tiempo breve.
Figura 32
4.1.4 Preselección del volumen de la señal de sonido
Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón
"Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) destella la palabra "Service". Ahora seleccione la letra "U" con
la perilla de ajuste (1). Como resultado, en la pantalla "RPM | RCF" (A-1) aparecen las palabras "Vol=0-
9/Sound". Después de presionar el botón "RPM | RCF" (4), puede ajustar el volumen deseado entre 0 (bajo) y 9
(alto) con la perilla de ajuste (1) (vea la Figura 33).
Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa
normal al apagar la centfuga durante un tiempo breve.
Figura 33
4.1.5 Selección de meloa para la señal del sonido, final de operación
Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón
"Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) destella la palabra "Service". Ahora seleccione la letra "G"
con la perilla de ajuste (1). Como resultado, en la pantalla "RPM | RCF" (A-1) aparece la palabra "SonGo/Sound".
Después de presionar el botón "RPM | RCF" (4), puede seleccionar una melodía con la perilla de ajuste (1). (Vea
la Figura 34.)
Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa
normal al apagar la centrífuga durante un tiempo breve.
ES-16 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figura 34
4.1.6 Encendido y apagado del sonido del teclado
Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón
"Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) destella la palabra "Service". Ahora seleccione la letra "B" con
la perilla de ajuste (1). Como resultado, en la pantalla "RPM | RCF" (A-1) aparece la palabra "ON/BEEP ".
Después de presionar el botón "RPM | RCF" (4), puede encender el sonido del teclado (On) o apagarlo (Off) con
la perilla de ajuste (1). (Vea la Figura 38.)
Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa
normal al apagar la centrífuga durante un tiempo breve.
Figura 35
4.1.7 Extracción de datos operativos
ATENCIÓN:
Esto solo debe realizarlo un usuario avanzado o ingeniero de servicio.
En el modo "Basic Adjustments" puede extraer los datos operativos de la centfuga. Proceda como
se describe en el punto 4.1.2 para ingresar en este modo del programa. Presione el botón "Accel/Decel" (5). En la
pantalla "Acc/Dec" (A-2) destella la palabra "Service".
El acceso a la diferente información se logra con la perilla de ajuste (1):
A= arranques previos de la centfuga
H= horas de operación previa
S= versión del software
r= software del convertidor
E= lista de mensajes de error previos
h= tiempo de operación del motor
La lista de los últimos 99 mensajes de error puede consultarse al presionar el botón "RPM | RCF" (4) y
desplazarse en ella con la perilla de ajuste (1). Los códigos de error respectivos aparecen en la pantalla "RPM |
RCF" (A-1). Consulte la "Tabla 6: mensajes de error" (vea APÉNDICE).
Para regresar al modo de programa normal, apague la centrífuga durante un tiempo breve.
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-17
Figura 36
5. MANTENIMIENTO
5.1 Mantenimiento y limpieza
5.1.1 Cuidado general:
El mantenimiento de la centfuga se limita a mantener el rotor, la cámara del rotor y los accesorios del rotor
limpios, así como lubricar regularmente los pernos de insercn del rotor de un rotor basculante (si es disponible).
El lubricante más indicado es el aceite High TEF.
No se permite el uso de lubricantes que contienen molycote y grafito.
Preste atención especial a las partes de aluminio anodizado. La ruptura de los rotores puede ser causada incluso
por el daño más ligero.
En el caso de los rotores, portatubos o racks de tubos que entren en contacto con sustancias corrosivas, los
puntos en cuestión deben limpiarse con cuidado.
Algunos ejemplos de sustancias corrosivas son: álcalis, soluciones de jabón alcalinas, aminas alcalinas, ácidos
concentrados, soluciones que contienen metales pesados, solventes clorados sin agua, soluciones salinas como
agua con sal, fenol e hidrocarburos halogenados.
Limpieza de unidades, rotores y accesorios:
- Apague el dispositivo y desconéctelo de la corriente eléctrica antes de comenzar la limpieza o desinfección. No
vierta quidos en el interior de la caja.
- No rocíe desinfectante sobre el dispositivo.
- La limpieza profusa no solo tiene el objetivo sanitario sino también el de evitar la corrosión debido a la
contaminación.
- Para evitar el daño a las partes anodizadas como rotores, placas de reducción, etc., solo se deben usar
detergentes de pH neutro con un pH de 6 a 8. No deben usarse agentes de limpieza alcalinos (pH > 8).
- Después de la limpieza, asegúrese de secar todas las partes completamente, ya sea a mano o en un gabinete de
aire caliente (temperatura máxima de +50 °C).
- Es necesario recubrir las partes de aluminio anodizado con aceite contra la corrosión regularmente a fin de
incrementar la vida útil y reducir la predisposición a la corrosión.
- Debido a la humedad o a las muestras no selladas herméticamente, se puede formar condensado. El
condensado debe eliminarse de la cámara del rotor con una tela suave regularmente.
El procedimiento de mantenimiento debe repetirse cada 10 a 15 operaciones, o al menos una vez a la
semana.
- Conecte la unidad a la corriente eléctrica hasta después de que el equipo esté completamente seco.
- No realice desinfección con rayos UV, beta o gamma ni otro tipo de radiación de enera.
- Los rotores de metal pueden usarse en autoclave.
- La tapa del rotor y adaptadores también pueden usarse en autoclave (máximo 121 °C, 20 minutos).
- Los racks para tubos están hechas de PP y no pueden usarse en autoclave a 134 °C.
ES-18 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
5.1.2 Limpieza y desinfección de la unidad
1. Abra la tapa antes de apagar la unidad. Descoctela de la alimentación eléctrica.
2. Abra la tuerca del rotor al girar la llave del rotor en sentido contrario al reloj.
3. Retire el rotor.
4. Para la limpieza y desinfección de la unidad y la cámara del rotor, use el limpiador mencionado anteriormente.
5. Limpie todas las áreas accesibles del dispositivo y sus accesorios, incluyendo el corn eléctrico, con una tela
humedecida.
6. Lave los sellos de caucho y la cámara del rotor completamente con agua.
7. Frote los sellos de caucho con glicerol o talco para prevenir que se hagan quebradizos. Otros componentes de
la unidad, como el seguro de la tapa, el eje del motor y el rotor no deben engrasarse.
8. Seque el eje del motor con una tela suave, seca y sin pelusa.
9. Controle la unidad y accesorios por posible do.
Aserese de que la centrífuga esté apagada y desconéctela de la corriente eléctrica. Luego remueva el polvo
adherido de las ranuras de ventilación en la centrífuga con un cepillo suave. Haga esto al menos cada seis meses.
5.1.3 Limpieza y desinfección del rotor
1. Limpie y desinfecte los rotores, tapas de los rotores y adaptadores con el limpiador mencionado anteriormente.
2. Use un cepillo para botellas para limpiar y desinfectar los orificios internos del rotor.
3. Enjuague los rotores, tapa del rotor y adaptador con agua limpia. En particular, las perforaciones de los rotores
angulares.
4. Para secar los rotores y accesorios, colóquelos en una toalla. Coloque los rotores angulares con los orificios
hacia abajo.
5. Seque el cono del rotor con una tela suave, seca y sin pelusa e inspeccione por posible daño. No engrase el
cono del rotor.
5.1.4 Desinfección de rotores de aluminio
En caso de que material infeccioso se derrame dentro de la centrífuga, el rotor y cámara del rotor deben
desinfectarse inmediatamente después de la operación. Los rotores pueden esterilizarse en autoclave a una
temperatura máxima de 121 °C.
5.1.5 Desinfección de rotores de PP
Uso en autoclave
Tiempo recomendado en autoclave: 15 a 20 minutos a 121 °C (1 bar)
ATENCIÓN:
No debe excederse el tiempo de esterilización de 20 minutos. La esterilización repetida causará
reducción de la resistencia mecánica del material plástico.
Antes de introducirse en autoclave, el rotor de PP y adaptador deben limpiarse completamente para
evitar que los residuos de suciedad se quemen dentro.
Puede hacer caso omiso de las consecuencias de algunos residuos químicos en materiales plásticos a
temperaturas ambiente. Sin embargo, a altas temperaturas durante la esterilización en autoclave, los residuos
pueden corroer y destruir el plástico. Los objetos deben enjuagarse completamente con agua destilada después
de limpiarse pero antes de la esterilización. Los residuos de cualquier líquido de limpieza pueden causar fisuras,
blanqueamiento y manchas.
Esterilización con gas
Los adaptadores, botellas y rotores pueden esterilizarse con óxido de etileno. Asegúrese de ventilar los artículos
después de la esterilización y antes de volver a usarlos.
ATENCIÓN:
Puesto que la temperatura puede aumentar durante la esterilización, los rotores, adaptadores y botellas
no deben cerrarse y deben destornillarse completamente
Esterilización química
Las botellas, adaptadores y rotores pueden tratarse con los desinfectantesquidos usuales.
ATENCIÓN:
Antes de aplicar cualquier otro método de limpieza o descontaminación que no sea el recomendado por
el fabricante, ngase en contacto con este para asegurarse de que no daña la unidad o el rotor.
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-19
5.1.6 Ruptura de vidrio
Con altos valores g, el porcentaje de ruptura de tubos de vidrio se incrementa. Las astillas de vidrio deben retirarse
inmediatamente del rotor, portatubos, adaptadores y de la cámara del rotor misma. Las astillas de vidrio finas
rayarán y por lo tanto dañarán la cubierta superficial protectora de un rotor. Si permanecen astillas en la cámara
del rotor, se acumulará polvo fino de metal debido a la circulación del aire. Este polvo metálico negro y muy fino
contamina significativamente la cámara del rotor, el rotor, los portatubos y las muestras.
Si es necesario, remplace los adaptadores, tubos y accesorios para evitar más daños. Inspeccione regularmente
los orificios internos del rotor por posibles residuos y daño.
ATENCIÓN:
Verifique las especificaciones relevantes de las centfugas de tubos con el fabricante.
5.2 Vida útil de rotores, portatubos y accesorios
Los rotores y tapa del rotor hechos de aluminio o acero inoxidable tienen una vida útil máxima de siete os a
partir del primer uso.
Las tapas y tapones transparentes de rotor hechos de PC o PP, así como los rotores, racks de tubos y
adaptadores de PP tienen un tiempo operativo máximo de tres os a partir del primer uso.
Condición para el tiempo operativo: condición de uso adecuado libre de daños, con el cuidado recomendado.
6. IDENTIFICACIÓN Y SOLUCIÓN DE PROBLEMAS
6.1 Mensaje de error: Causa / Solución
Los mensajes de error se presentan para ayudar a localizar posibles errores más rápido.
El diagnóstico al que se hace referencia en este catulo puede no siempre ser el caso, ya que son solo errores y
soluciones que ocurren teóricamente.
Le pedimos que nos mantenga informados acerca de cualquier tipo de error que ocurra y que no esté descrito en
este catulo. Solo mediante su información podremos mejorar este manual de operación.
Muchas gracias de antemano por su apoyo.
6.2 Estudio de posibles mensajes de error y sus soluciones
6.2.1 Apertura de la tapa durante una falla de la corriente eléctrica (Apertura de emergencia de la tapa)
En caso de falla de la corriente eléctrica o mal funcionamiento, la tapa de la centrífuga puede abrirse manualmente
para recuperar las muestras.
En los modelos FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (cerradura accionada por motor)
Proceda de la siguiente manera:
1. Apague la centrífuga y desenchufe el corn eléctrico y espere hasta que el rotor se haya detenido
completamente (esto puede durar varios minutos).
2. En el lado izquierdo de la caja de la centrífuga hay un detenedor de plástico. Retire este detenedor y detrás de
él hay una tuerca hexagonal.
3. Tome la llave incluida, colóquela en el agujero y bloquee la llave con la tuerca hexagonal (vea la Figura 37).
4. Ahora gire la llave a la derecha (en sentido del reloj) hasta el límite.
ATENCIÓN:
a) Solo gire hasta el límite; no apriete la tuerca.
b) Ahora abra la tapa de la centrífuga.
c) Encienda la centrífuga nuevamente para reanudar
el trabajo.
ES-20 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figura 37
For 5714
Realice lo siguiente (ver la Figura 38):
ATENCIÓN:
Apague la centfuga y desconecte el cable de alimentación, espere hasta que el rotor se
detenga completamente (esto puede demorar varios minutos)
En el lado derecho de la centrífuga hay un detenedor plástico (Figura 38). Quite este detenedor,
que está conectado a la cerradura de la tapa, horizontalmente desde la carcasa hasta que la tapa
de la centrífuga se abra.
Ahora abra la tapa de la centrífuga
Figura 38
6.2.2 Descripción del sistema de mensajes de error
El mensaje de error "error" (M11) se muestra en la pantalla "Time/Prog" (A-3) (vea la Figura 39). La información
detallada acerca de los posibles mensajes de error está en la "tabla 6: mensajes de error" (vea Andice).
Figura 39
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-21
7. RECEPCIÓN DE CENTRÍFUGAS PARA REPARACIÓN
Riesgo para la salud por equipo, rotores y accesorios contaminados.
En caso de enviar la centfuga para reparación al fabricante, observe lo siguiente:
La centrífuga debe descontaminarse y limpiarse antes del envío para proteccn de las personas,
medio ambiente y material.
Certificado de descontaminación en la entrega de devolución de bienes (vea APÉNDICE)
Nos reservamos el derecho de no aceptar centrífugas contaminadas.
Todos los costos adicionales incurridos para la limpieza y desinfección de las unidades s e agregarán al débito de
la cuenta del cliente
8. TRANSPORTE, ALMACENAMIENTO Y ELIMINACIÓN
8.1 Transporte
Antes del transporte, retire el rotor.
Solo transporte la unidad en el embalaje original.
Use un auxiliar de transporte para el transporte a través de largas distancias para fijar el eje del motor.
Temperatura del aire
Humedad relativa
Presión de aire
Transporte general
-25 a 60 °
10 a 75 %
30 a 106 kPa
8.2 Almacenamiento
Durante el almacenamiento de la centrífuga, se deben observar las siguientes
condiciones ambientales:
Temperatura del aire
Humedad relativa
Presión de aire
En empaquetado de
transporte
-25 a 55 °
10 a 75 %
70 a 106 kPa
8.3 Transporte, instalación, traslado y desecho de la centrífuga
FC5714/FC5718/FC5718R/FC5816/FC5816R
Estas instrucciones complementan las anteriores instrucciones generales del capítulo 8 pero no las sustituyen.
8.3.1 Transporte
Transporte el dispositivo en su embalaje original
Dos personas deben trasportar siempre la centrífuga.
Para largas distancias, utilice la ayuda de un transporte.
8.3.2 Instalación
Dos personas deben trasportar siempre la centfuga.
Abrir la caja y sacar el dispositivo.
1. Corte la cinta adhesiva.
2. Abra las 4 solapas de la caja.
3. Saque los accesorios.
4. Introduzca las manos hasta colocarlas debajo del dispositivo y, con
ayuda de otra persona, saque la centrífuga de la caja.
ES-22 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
ADVERTENCIA: Peligro de levantamiento. Existe riesgo de lesiones si una sola persona la
levanta. Solicite ayuda al levantar o mover el equipo.
Colocar el dispositivo en una mesa de laboratorio estable, horizontal y no
resonante.
1. Retire el material protector para el transporte delantero y trasero.
2. Retire la funda de plástico.
3. Mantenga una distancia nima de 30 cm entre los laterales del equipo y los
dispositivos adyacentes y entre la parte trasera y la pared.
4. Instale el dispositivo en una ubicación con buena ventilación que esté
protegida del sol directo para evitar que se recaliente.
Conectar el dispositivo
1. Después de realizar la instalación, espere durante cuatro horas para encender
la centrífuga para evitar dañar el compresor.
2. Compruebe que la tensión y la frecuencia de la red eléctrica se ajustan a los
requisitos de la placa de características del dispositivo (consulte la parte
trasera del dispositivo) y después conecte el dispositivo a la red eléctrica.
Retirar el material protector para el transporte de la cámara del rotor.
1. Encienda el dispositivo con el interruptor de encendido.
2. Abra la tapa de la centrífuga utilizando el botón de apertura.
3. Retire el material protector para el transporte.
4. Coloque el rotor de forma vertical en el eje del motor.
5. Gire la tuerca del rotor con la llave incluida en el sentido de las agujas del reloj hasta que quede apretada.
Ya puede utilizar el dispositivo.
Conserve el embalaje y todo el material protector para el transporte para mover el dispositivo en un futuro.
8.3.3 Embalaje
Embale la centrífuga siguiendo los pasos anteriores en el sentido contrario.
8.3.4 Entrega del dispositivo
Al pasar el equipo a terceros, por favor asegúrese de incluir también este manual
de instrucciones.
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-23
9. FICHA TÉCNICA
9.1 Especificaciones
9.1.1 Centrifuge FC5714
Modelo
FC5714
Rango de v elocidad
200 rpm -14000 rpm;10 rpm/set
Máxima RCF
18624 x g;10 x g/set
Máxima capacidad (rotor)
4x100ml
Rango de temperatura
Enfriamiento por aire
Tiempo de operación
10 seg a 99 hr 59 min 59 seg o continuo
Niv el de ruido (dependiendo del
rotor)
63 ± 2 dB(A)
Densidad permitida a xima
v elocidad
1.2 g/ml
Energía cinética permitida
5595 Nm
Conexión CA de corriente
ectrica principal
230 V ~ 50/60 Hz
230 V ~ 50/60
Hz
Fluctuación de v oltaj e
± 10 %
± 10 %
Consumo de corriente
1.3 A
1.3 A
Consumo de energía
240 W
240 W
Dimensiones (Ancho ×
Profundidad × Alto)
362 x 493 x 330 mm
Peso neto (sin rotor)
14.3 x 19.4 x 13.0 in
Dimensiones de envío (Ancho ×
Profundidad x Alto)
30 kg
Peso de env ío (sin rotor)
66 lb
Condiciones ambientales (EN/IEC 61010-1)
Ambiente
para uso interno solamente
Altitud
Use hasta a una altitud de 2000 m
Temperatura ambiente
2 °C hasta 35 °C
Máx. humedad relativ a
Máx. humedad relativa 80 % para temperaturas de
hasta 31 °C,
disminuyendo linealmente a 50% de humedad relativa
hasta 35 °C.
Categoría de sobrev oltaje
(IEC 60364-4-443)
II
Grado de contaminación
2
Clase de protección
I
No adecuada para usarse en ambientes peligrosos.
ES-24 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.2 Centfuga FC5718
Modelo
FC5718
Rango de v elocidad
200 rpm -18000 rpm;10 rpm/set
Máxima RCF
23542 x g;10 x g/set
Máxima capacidad (rotor)
4x100ml
Rango de temperatura
Enfriamiento por aire
Tiempo de operación
10 seg a 99 hr 59 min 59 seg o continuo
Niv el de ruido (dependiendo del
rotor)
60 ± 2 dB(A)
Densidad permitida a xima
v elocidad
1.2 g/ml
Energía cinética permitida
16672 Nm
Conexión CA de corriente
ectrica principal
230 V ~ 50/60 Hz
230 V ~ 50/60 Hz
Fluctuación de v oltaj e
± 10 %
± 10 %
Consumo de corriente
2.0 A
2.0 A
Consumo de energía
455 W
455 W
Dimensiones (Ancho ×
Profundidad × Alto)
408 x 499 x 351 mm
Peso neto (sin rotor)
16.1 x 19.7 x 13.8 in
Dimensiones de envío (Ancho ×
Profundidad x Alto)
43 kg
Peso de env ío (sin rotor)
95 lb
Condiciones ambientales (EN/IEC 61010-1)
Ambiente
para uso interno solamente
Altitud
Use hasta a una altitud de 2000 m
Temperatura ambiente
2 °C hasta 35 °C
Máx. humedad relativ a
Máx. humedad relativa 80 % para temperaturas de
hasta 31 °C,
disminuyendo linealmente a 50% de humedad relativa
hasta 35 °C.
Categoría de sobrev oltaje
(IEC 60364-4-443)
II
Grado de contaminación
2
Clase de protección
I
No adecuada para usarse en ambientes peligrosos.
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-25
9.1.3 Centrifuge FC5718R
Modelo
FC5718R
Rango de v elocidad
200 rpm -18000 rpm;10 rpm/set
Máxima RCF
23542 x g;10 x g/set
Máxima capacidad (rotor)
4x100ml
Rango de temperatura
-20~40
Tiempo de operación
10 seg a 99 hr 59 min 59 seg o continuo
Niv el de ruido (dependiendo del
rotor)
60 ± 2 dB(A)
Densidad permitida a xima
v elocidad
1.2 g/ml
Energía cinética permitida
25111 Nm
Conexión CA de corriente
ectrica principal
230 V ~ 50/60 Hz
230 V ~ 50/60 Hz
Fluctuación de v oltaj e
± 10 %
± 10 %
Consumo de corriente
3.0 A
3.0 A
Consumo de energía
660 W
660 W
Dimensiones (Ancho ×
Profundidad × Alto)
407 x 731 x 359 mm
Peso neto (sin rotor)
16.0 x 28.8 x 14.1 in
Dimensiones de envío (Ancho ×
Profundidad x Alto)
60 kg
Peso de env ío (sin rotor)
132 lb
Condiciones ambientales (EN/IEC 61010-1)
Ambiente
para uso interno solamente
Altitud
Use hasta a una altitud de 2000 m
Temperatura ambiente
2 °C hasta 35 °C
Máx. humedad relativ a
Máx. humedad relativa 80 % para temperaturas de
hasta 31 °C,
disminuyendo linealmente a 50% de humedad relativa
hasta 35 °C.
Categoría de sobrev oltaje
(IEC 60364-4-443)
II
Grado de contaminación
2
Clase de protección
I
No adecuada para usarse en ambientes peligrosos.
ES-26 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.4 Centrifuge FC5816
Modelo
FC5816
Rango de v elocidad
200 rpm -15000 rpm;10 rpm/set
Máxima RCF
21379 x g;10 x g/set
Máxima capacidad (rotor)
6 x 250ml
Rango de temperatura
Enfriamiento por aire
Tiempo de operación
10 seg a 99 hr 59 min 59 seg o continuo
Niv el de ruido (dependiendo del
rotor)
61 ± 2 dB(A)
Densidad permitida a xima
v elocidad
1.2 g/ml
Energía cinética permitida
34363 Nm
Conexión CA de corriente
ectrica principal
230 V ~ 50/60 Hz
230 V ~ 50/60 Hz
Fluctuación de v oltaj e
± 10 %
± 10 %
Consumo de corriente
2.4 A
2.4 A
Consumo de energía
530 W
530 W
Dimensiones (Ancho ×
Profundidad × Alto)
446 x 538 x 354 mm
Peso neto (sin rotor)
17.6 x 21.2 x 13.9 in
Dimensiones de env ío (Ancho ×
Profundidad x Alto)
52 kg
Peso de env ío (sin rotor)
115 lb
Condiciones ambientales (EN/IEC 61010-1)
Ambiente
para uso interno solamente
Altitud
Use hasta a una altitud de 2000 m
Temperatura ambiente
2 °C hasta 35 °C
Máx. humedad relativ a
Máx. humedad relativa 80 % para temperaturas de
hasta 31 °C,
disminuyendo linealmente a 50% de humedad relativa
hasta 35 °C.
Categoría de sobrev oltaje
(IEC 60364-4-443)
II
Grado de contaminación
2
Clase de protección
I
No adecuada para usarse en ambientes peligrosos.
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-27
9.1.5 Centrifuge FC5816R
Modelo
FC5816R
Rango de v elocidad
200 rpm -16000 rpm;10 rpm/set
Máxima RCF
24325 x g;10 x g/set
Máxima capacidad (rotor)
6 x 250ml
Rango de temperatura
-20~40
Tiempo de operación
10 seg a 99 hr 59 min 59 seg o continuo
Niv el de ruido (dependiendo del
rotor)
63 ± 2 dB(A)
Densidad permitida a xima
v elocidad
1.2 g/ml
Energía cinética permitida
34363 Nm
Conexión CA de corriente
ectrica principal
230 V ~ 50/60 Hz
230 V ~ 50/60 Hz
Fluctuación de v oltaj e
± 10 %
± 10 %
Consumo de corriente
3.7 A
3.7 A
Consumo de energía
785 W
785 W
Dimensiones (Ancho ×
Profundidad × Alto)
723 x 538 x 354 mm
Peso neto (sin rotor)
28.5 x 21.2 x 13.9 in
Dimensiones de envío (Ancho ×
Profundidad x Alto)
77 kg
Peso de env ío (sin rotor)
170 lb
Condiciones ambientales (EN/IEC 61010-1)
Ambiente
para uso interno solamente
Altitud
Use hasta a una altitud de 2000 m
Temperatura ambiente
2 °C hasta 35 °C
Máx. humedad relativ a
Máx. humedad relativa 80 % para temperaturas de
hasta 31 °C,
disminuyendo linealmente a 50% de humedad relativa
hasta 35 °C.
Categoría de sobrev oltaje
(IEC 60364-4-443)
II
Grado de contaminación
2
Clase de protección
I
No adecuada para usarse en ambientes peligrosos.
ES-28 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.6 Centrifuge FC5916
Model
FC5916
Speed Range
200 rpm -16000 rpm;10 rpm/set
Maximum RCF
24325 x g;10 x g/set
Maximum Capacity(Rotor)
4x750ml
Temperature range(N/A)
Air cool
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise lev el (depending on the
rotor)
63 ± 2 dB(A)
Allowable density at maximum
speed
1.2 g/ml
Allowable kinetic energy
60629 Nm
Mains power connection AC
230 V ~
50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
2.8 A
5.6 A
Power consumption
640 W
680 W
Dimensions (W × D × H)
540 x 670 x 390 mm
21.3 x 26.4 x 15.4 in
Net Weight (without rotor)
85 kg
187 lb
Shipping Dimensions (W × D × H)
780 x 670 x 590 mm
30.7 x 26.4 x 23.3 in
Shipping Weight (without rotor)
98 kg
216 lb
Ambient conditions (EN/IEC 61010-1)
Environment
For indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
2°C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to
31°C,
decreasing linearly to 50 % relative humidity up to
35°C.
Ov ervoltage category (IEC 60364-4-
443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-29
9.1.7 Centrifuge FC5916R
Model
FC5916R
Speed Range
200 rpm -16000 rpm;10 rpm/set
Maximum RCF
26331 x g;10 x g/set
Maximum Capacity(Rotor)
4x750ml
Temperature range(Digital)
-20~40
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise lev el (depending on the
rotor)
63 ± 2 dB(A)
Allowable density at maximum
speed
1.2 g/ml
Allowable kinetic energy
54458 Nm
Mains power connection AC
230 V ~
50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
7.2 A
20 A
Power consumption
1630 W
1750 W
Dimensions (W × D × H)
730 x 670 x 390 mm
28.7 x 26.4 x 15.4 in
Net Weight (without rotor)
118 kg
260 lb
Shipping Dimensions (W × D × H)
900 x 750 x 560 mm
40.0 x 29.5 x 22.0 in
Shipping Weight (without rotor)
137 kg
137 kg
Ambient conditions (EN/IEC 61010-1)
Environment
for indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
2°C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to
31°C,
decreasing linearly to 50 % relative humidity up to
35°C.
Ov ervoltage category (IEC 60364-4-
443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
ES-30 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.2 Diagramas y dimensiones
D
H
W
W
H
D
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-31
Modelo
W (mm / in.)
D (mm / in.)
H (mm / in.)
FC5714
362 / 14.3
493 / 19.4
330 / 13.0
FC5718
408 / 16.1
499 / 19.7
351 / 13.8
FC5718R
408 / 16.1
731 / 28.8
359 / 14.1
FC5816
446 / 17.6
538 / 21.2
354 / 13.9
FC5816R
723 / 28.5
538 / 21.2
354 / 13.9
FC5916
540 / 21.3
670 / 26.4
390 / 15.4
FC5916R
730 / 28.7
670 / 26.4
390 / 15.4
10. CONFORMIDAD
La marca correspondiente en el producto indica la conformidad con los siguientes estándares.
Marca
Esndar
Este producto es conforme con las normas armonizadas aplicables de directivas
de la UE 2011/65/UE (RoHS) 2014/30/UE (EMC) y 2014/35/UE (LVD).
La declaración de conformidad de la UE está disponible
en línea en www.ohaus.com/ce.
Desecho
En conformidad con la Directiva Europea 2012/19/UE sobre Residuos de
Equipos Eléctricos y Electnicos (WEEE), este dispositivo no puede
desecharse con la basura doméstica. Esto también es aplicable para países
fuera de la UE, según sus requerimientos específicos.
Deseche este producto de acuerdo con las regulaciones locales en el punto de
recolección especificado para equipos eléctricos y electrónicos.
Si tiene alguna pregunta, comuníquese con la autoridad responsable o con el
distribuidor a quien compró este dispositivo.
En caso de que este dispositivo sea transferido a otras partes (para uso privado
o profesional), también deberá mencionarse el contenido de esta regulación.
Para obtener instrucciones de eliminación en Europa, consulte
www.ohaus.com/weee.
Gracias por su contribución a la protección ambiental.
ES-32 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11. APÉNDICE
TABLA 1: PESO NETO PERMITIDO
TABLA 2: TEMPERATURAS MÍNIMAS A XIMA VELOCIDAD
TABLA 3: MÁXIMA VELOCIDAD Y VALORES RCF PARA ROTORES PERMITIDOS
TABLA 4: TIEMPOS DE ACELERACIÓ N Y DESACELERACIÓ N
TABLA 5: MENSAJES DE ERROR
TABLA 6 CORRECCIÓ N DEL RADIO
TABLA 7: FORMULARIO DE DEVOLUCIÓN / CERTIFICADO DE DESCONTAMINACIÓ N
11.1 Table 1: PESO NETO PERMITIDO
Nú mero del rotor
Descripció n del rotor
Peso neto
permitido
Velocidad má x. ( x1000 rpm)
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
16.0
16.0
30314820
Rotor, Sw ing out, 4x250ml, ID
4x 536 g
4.5
4.5
30314821
Rotor, Angle, 6x250ml FB, ID
6x 355 g
8.0
8.0
8.0
8.0
30314822
Rotor, Sw ing out, 4x 100ml, ID
4x 340 g
4.5
4.5
4.5
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
4x 465 g
4.0
5.0
5.0
30314824
Rotor, Sw ing out, 2x3MTP, ID
2x 310 g
4.5
4.5
4.5
4.5
4.5
4.5
4.5
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
6x 140 g
11.0
13.5
30314826
Rotor, Angle, 6x85ml RB, ID
6x 140 g
9.0
9.0
11.0
13.0
11.0
11.0
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
4x 140 g
12.0
12.0
12.0
12.0
15.0
15.0
30314828
Rotor, Sw ing out, 16x50ml, ID
4x 557 g
4.5
4.5
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-33
30314829
Rotor, Angle, 10x50ml FA, ID
10x 74 g
7.5
7.5
9.0
10.5
10.0
10.0
30314830
Rotor, Angle, 6x50ml RB/FA, ID
6x 72 g
6.0
6.0
6.0
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
6x 94 g
12.0
12.0
13.0
13.0
13.0
13.0
30314832
Rotor, Angle, 30x15ml RB/FA, ID
30x 32 g
4.5
4.5
4.5
4.5
4.5
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
20x 18 g
12.0
12.0
12.0
12.0
12.0
12.0
30314834
Rotor, Angle, 12x15ml RB/FA, ID
12x 25 g
6.0
6.0
6.0
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
44x 3.8 g
13.5
13.5
15.0
16.0
16.0
16.0
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
30x 3.4 g
12.0
13.0
14.0
15.0
15.0
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
15.0
15.0
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
12x 3.4 g
18.0
18.0
30314840
Rotor, Angle, 64x0.5ml, ID
64x 2.3 g
13.5
13.5
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
4x 3.5 g
15.0
15.0
15.0
15.0
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
4x 995 g
4.0
4.0
30553086
Rotor Angle 4x 500ml D69mm ID
4x 708 g
8.0
8.0
ES-34 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.2 Table 2: TEMPERATURAS NIMAS A XIMA VELOCIDAD
Nú mero de rotor
Descripció n del rotor
Modelo
Velocidad
má xima
N-max
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314821
Rotor, Angle, 6x250ml FB, ID
FC5916R
8000 rpm
-1 ° C
30314822
Rotor, Sw ing out, 4x 100ml, ID
FC5718R
4500 rpm
-7 ° C
FC5816R
4500 rpm
-7 ° C
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5718R
5000 rpm
-3 ° C
FC5816R
5000 rpm
-3 ° C
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5718R
4500 rpm
-6 ° C
FC5816R
4500 rpm
-6 °C
FC5916R
4500 rpm
-8 ° C
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718R
13500 rpm
17 ° C
FC5816R
13500 rpm
17 ° C
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718R
9000 rpm
-5 ° C
FC5816R
9000 rpm
-5 ° C
FC5916R
13000 rpm
8 ° C
30314827
Rotor, Angle, 4x 85ml RB, ID, Hi
FC5718R
12000 rpm
5 ° C
FC5816R
12000 rpm
5 ° C
FC5916R
16000 rpm
7 ° C
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718R
7500 rpm
9 ° C
FC5816R
7500 rpm
9 ° C
FC5916R
10500 rpm
2 ° C
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5718R
6000 rpm
-8 ° C
FC5816R
6000 rpm
-8 ° C
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718R
12000 rpm
-5 ° C
FC5816R
12000 rpm
-5 ° C
FC5916R
13000 rpm
-4 ° C
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5718R
4500 rpm
-9 ° C
FC5816R
4500 rpm
-9 °C
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718R
12000 rpm
2 ° C
FC5816R
12000 rpm
2 ° C
FC5916R
12000 rpm
-2 ° C
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5718R
6000 rpm
-11.5 ° C
FC5816R
6000 rpm
-11.5 ° C
30314835
Rotor, Angle, 44x 1.5/2.0ml, ID
FC5718R
13500 rpm
1 ° C
FC5816R
13500 rpm
1 ° C
FC5916R
16000 rpm
3 ° C
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5718R
14000 rpm
9 ° C
FC5816R
14000 rpm
9 ° C
FC5916R
15000 rpm
3 ° C
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718R
18000 rpm
-2 ° C
FC5816R
18000 rpm
-2 ° C
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718R
13500 rpm
2 ° C
FC5816R
13500 rpm
2 ° C
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718R
15000 rpm
-1 ° C
FC5816R
15000 rpm
-1 ° C
FC5916R
15000 rpm
-2 ° C
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916R
4500 rpm
4 ° C
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916R
8000 rpm
0 ° C
Todas las indicaciones de temperatura se refieren a una temperatura ambiente de 23 ° C. Al exceder
este valor o la radiación solar directa a la centrífuga, estos valores no pueden mantenerse.
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-35
11.3 Table 3: MÁXIMA VELOCIDAD Y VALORES RCF PARA ROTORES PERMITIDOS
Nú mero del rotor
Descripció n del rotor
Modelo
Velocidad má x.
RPM
RCF
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
4500
3780 x g
FC5816R
4500
3780 x g
30314821
Rotor, Angle, 6x250ml FB, ID
FC5816
8000
10016 x g
FC5816R
8000
10016 x g
FC5916
8000
10016 x g
FC5916R
8000
10016 x g
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5714
4500
3350 x g
FC5718
4500
3350 x g
FC5718R
4500
3350 x g
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
4000
2486 x g
FC5718
5000
3885 x g
FC5718R
5000
3885 x g
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5714
4500
2716 x g
FC5718
4500
2716 x g
FC5718R
4500
2716 x g
FC5816
4500
2716 x g
FC5816R
4500
2716 x g
FC5916
4500
2716 x g
FC5916R
4500
2716 x g
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
11000
13932 x g
FC5718R
13500
20984 x g
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
9000
10413 x g
FC5718R
9000
10413 x g
FC5816
11000
15555 x g
FC5816R
13000
21726 x g
FC5916
11000
15555 x g
FC5916R
13000
21726 x g
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718
12000
14809 x g
FC5718R
12000
14809 x g
FC5816
12000
14809 x g
FC5816R
12000
14809 x g
FC5916
15000
23140 x g
FC5916R
16000
26328 x g
30314828
Rotor, Sw ing out, 16x50ml, ID
FC5816
4500
3735 x g
FC5816R
4500
3735 x g
ES-36 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314829
Rotor, Angle, 10x 50ml FA, ID
FC5718
7500
8174 x g
FC5718R
7500
8174 x g
FC5816
9000
11771 x g
FC5816R
10500
16022 x g
FC5916
10000
14532 x g
FC5916R
10500
16022 x g
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
12000
13522 x g
FC5718R
12000
13522 x g
FC5816
13000
15869 x g
FC5816R
13000
15869 x g
FC5916
13000
15869 x g
FC5916R
13000
15869 x g
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5714
4500
2830 x g
FC5718
4500
2830 x g
FC5718R
4500
2830 x g
FC5816
4500
2830 x g
FC5816R
4500
2830 x g
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
12000
15775 x g
FC5718R
12000
15775 x g
FC5816
12000
15775 x g
FC5816R
12000
15775 x g
FC5916
12000
15775 x g
FC5916R
12000
15775 x g
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718
13500
17113 x g
FC5718R
13500
17113 x g
FC5816
15000
21128 x g
FC5816R
16000
24039 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5714
12000
15131 x g
FC5718
13000
17758 x g
FC5718R
14000
20595 x g
FC5916
15000
23643 x g
FC5916R
15000
23643 x g
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-37
30314839
Rotor, Angle, 12x 1.5/2.0ml, ID
FC5718
18000
23542 x g
FC5718R
18000
23542 x g
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
13500
16298 x g
FC5718R
13500
16298 x g
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
15000
15343 x g
FC5718R
15000
15343 x g
FC5916
15000
15343 x g
FC5916R
15000
15343 x g
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916
4000
3434 x g
FC5916R
4500
4346 x g
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916
8000
10367 x g
FC5916R
8000
10367 x g
ES-38 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.4 Table 4: TIEMPOS DE ACELERACIÓN Y DESACELERACIÓN
Nú mero de
producto
Descripció n del rotor
Modelo
Aceleració n* Tiempo en
seg.
Desaceleració n* Tiempo
en seg.
Nivel 0
Nivel 9
Nivel 0
Nivel 9
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5714
200
25
230
35
FC5718
240
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
20
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
309
34
458
36
FC5816R
309
34
458
36
30314821
Rotor, Angle, 6x250ml FB, ID
FC5816
664
130
2906
92
FC5816R
664
130
2906
92
FC5916
573
66
1903
84
FC5916R
573
66
1903
84
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5714
110
14
170
19
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
110
14
170
17
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5714
230
25
340
26
FC5718
160
20
360
15
FC5718R
160
20
360
15
FC5816
452
43
616
38
FC5816R
452
43
616
38
FC5916
249
27
488
23
FC5916R
249
27
488
23
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
400
40
960
40
FC5718R
500
60
1260
50
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
360
40
1050
40
FC5718R
360
40
1050
40
FC5816
697
85
2313
70
FC5816R
2313
70
1630
76
FC5916
463
48
1654
46
FC5916R
549
69
1307
67
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718
300
60
820
40
FC5718R
300
60
820
40
FC5816
506
60
1745
49
FC5816R
506
60
1745
49
FC5916
448
50
1251
45
FC5916R
448
50
1251
45
30314828
Rotor, Sw ing out, 16x50ml, ID
FC5816
311
34
387
36
FC5816R
311
34
387
36
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718
380
63
796
37
FC5718R
380
62
1069
36
FC5816
753
115
2395
72
FC5816R
753
115
2395
72
FC5916
480
60
1747
68
FC5916R
480
60
1747
68
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5714
101
11
206
14
FC5718
89
13
239
11
FC5718R
90
12
463
11
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
446
48
1323
49
FC5816R
446
48
1323
49
FC5916
264
28
921
32
FC5916R
264
28
921
32
Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-39
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5714
157
19
370
20
FC5718
160
15
380
10
FC5718R
160
15
380
10
FC5816
149
25
985
20
FC5816R
149
25
985
20
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
512
54
1439
47
FC5816R
512
54
1439
47
FC5916
305
32
988
37
FC5916R
305
32
988
37
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
102
11
167
14
FC5718
88
13
222
11
FC5718R
88
12
433
11
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718
196
22
204
18
FC5718R
196
22
204
18
FC5816
257
29
436
28
FC5816R
274
31
485
29
FC5916
233
25
283
30
FC5916R
233
25
500
25
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5714
250
27
280
34
FC5718
200
25
360
25
FC5718R
210
25
360
30
FC5916
221
23
561
30
FC5916R
221
23
561
30
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5714
200
22
230
35
FC5718
210
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
30
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718
240
25
210
20
FC5718R
240
25
210
20
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
150
17
170
12
FC5718R
150
17
170
12
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
130
16
130
12
FC5718R
130
16
130
12
FC5916
100
12
201
12
FC5916R
100
12
201
12
30553085
Rotor SwingOut 4x750ml D99mm ID SL
FC5916
483
47
1287
49
FC5916R
551
55
1501
54
30553086
Rotor Angle 4x500ml D69mm ID
FC5916
575
73
2317
82
FC5916R
575
73
2317
82
* Nota: acelera de 0 a Vmax; desacelera de Vmax a 0.
ES-40 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.5 Tabla 5: Mensajes de error
Error-No.:
Descripció n
1
Aparece desbalance
2
Sensor de desbalance defectuoso
4
Interruptor de desbalance activado por má s de 5 segundos
8
Transpondedor en rotor defectuoso
11
Sensor de temperatura defectuoso
12
Exceso de temperatura de la cá mara
14
Salto de velocidad demasiado grande entre dos mediciones
Tapa CERRADA
33
Tapa abierta mientras el motor está funcionando
34
Contacto de la tapa defectuoso
38
Motor de la tapa bloqueado
40
Comunicació n con convertidor de frecuencia alterada durante el arranque
41
Comunicació n con convertidor de frecuencia alterada durante el paro
42
Cortocircuito en el convertidor de frecuencia
43
Bajo voltaje de convertidor de frecuencia
44
Alto voltaje del convertidor de frecuencia
45
Exceso de temperatura del convertidor de frecuencia
46
Exceso de temperatura del motor
47
Exceso de corriente del convertidor de frecuencia
48
Tiempo excedido entre la unidad de control y el controlador de frecuencia
49
Otro error del convertidor de frecuencia
55
Exceso de velocidad
70
Tiempo excedido entre el controlador y la interfaz RS232
99
Rotor no permitido en esta centrífuga
FALSO
El rotor insertado no existe en el programa
No. de rotor
Rotor no detectado
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-41
11.6 Table 6: CORRECCIÓN DEL RADIO
Rotor
Descripció n del
rotor
Contenedo
r/Adaptad
or
Descripció n del
Contenedor/Adaptador
Radio
(cm)
Correcció
n
(cm)
Compatibilidad
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle,
24x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314820
Rotor, Sw ing out,
4x 250ml, ID
Blank
NO Bucket/Adapter
16.7
30314903
Rack, 1x250ml D62mm FB, 2/pk
15.7
-1.0
30314904
Rack, 1x100ml D41mm RB, 2/pk
15.9
-0.8
30314905
Rack, 3x50ml D29mm RB, 2/pk
16.3
-0.4
30314906
Rack, 3x50ml D29mm FA, 2/pk
16.7
0.0
30314907
Rack, 1x50ml D34mm RB, 2/pk
15.8
-0.9
30314908
Rack, 2x 50ml D29mm Rim, 2/pk
15.8
-0.9
30314909
Rack, 1x50ml D29mm Rim, 2/pk
16.1
-0.6
30314910
Rack, 5x25ml D24.5mm RB, 2/pk
16.1
-0.6
30314911
Rack, 3x25ml D24.5mm Rim, 2/pk
15.5
-1.2
30314912
Rack, 9x15ml D17mm RB, 2/pk
16.3
-0.4
30314913
Rack, 7x15ml D17mm FA, 2/pk
16.4
-0.3
30314914
Rack, 7x15ml D17mm Rim, 2/pk
16.1
-0.6
30314915
Rack, 8x10ml D16mm RB, 2/pk
16.3
-0.4
30314916
Rack, 14x5-7ml D13mm RB, 2/pk
15.9
-0.8
30314917
Rack, 10x1.6-7ml D13mm Vac, 2/pk
15.9
-0.8
30304367
Rack, 8x4-10ml D16mm Vac, 2/pk
16.3
-0.4
30304368
Adapter, 9x 1.5/2.0ml D11mm, 2/pk
15.7
-1.0
30314821
Rotor, Angle,
6x 250ml FB, ID
Blank
NO Bucket/Adapter
14.1
30304369
Adapter, 8x 1.5ml D11mm, 2/pk
13.2
-0.9
30304370
Adapter, 5x 10ml D16mm RB, 2/pk
13.3
-0.8
30304371
Adapter, 4x 15ml D17mm FA, 2/pk
13.0
-1.1
30304372
Adapter, 2x 30ml D26mm RB, 2/pk
12.5
-1.6
30304373
Adapter, 1x 50ml D29mm RB, 2/pk
12.0
-2.1
30304374
Adapter, 1x 50ml D30mm FA, 2/pk
11.7
-2.4
30559412
Adapter 15x 1.5/2.0ml D11mm
MTBx 2 V1
9.5
13.6
30559413
Adapter 1x 250ml D60mm CBCx3 V1
11.8
-0.5
30559414
Adapter 1x 175/225ml D61mm FAx 2
V1
12.8
-2.3
30314822
Rotor, Sw ing out,
4x 100ml, ID
Blank
NO Bucket/Adapter
14.8
30314842
Rack, 1x100ml D46mm RB, 2/pk
14.1
-0.7
30314843
Rack, 1x85ml D38mm RB, 2/pk
14.1
-0.7
30314844
Rack, 1x50ml D34mm RB, 2/pk
14.2
-0.6
30314845
Rack, 1x50ml D29mm RB, 2/pk
14.2
-0.6
30314846
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.7
30314847
Rack, 1x 50ml D29mm Rim, 2/pk
14.3
-0.5
30314848
Rack, 1x30ml D25mm Rim, 2/pk
13.8
-1.0
30314849
Rack, 1x25ml D25mm RB, 2/pk
14.4
-0.4
30314850
Rack, 2x15ml D17mm FA, 2/pk
14.8
0.0
30314851
Rack, 1x15ml D17mm Rim, 2/pk
14.5
-0.3
-4.6
-0.5
ES-42 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314852
Rack, 7x15ml D17mm RB, 2/pk
14.5
-0.3
30314853
Rack, 9x5-7ml D13mm RB, 2/pk
14.2
-0.6
30314854
Rack, 10x1.5ml D11mm, 2/pk
9.9
-4.9
30314855
Rack, 9x4.5ml D12mm FB, 2/pk
14.0
-0.8
30314856
Rack, 7x5-7ml D13mm RB, 2/pk
11.6
-3.2
30314857
Rack, 7x5-7ml D13mm Vac, 2/pk
14.2
-0.6
30314858
Rack, 5x4-10ml D16mm Sar, 2/pk
14.3
-0.5
30314823
Rotor, Sw ing out,
4x 100ml, ID,
Sealable
Blank
NO Bucket/Adapter
14.6
30314860
Bucket, 100ml wo Cap, Sealable,
2/pk
14.2
-0.4
30314861
Bucket, 100ml w/ Cap, Sealable, 2/pk
14.2
-0.4
30314864
Rack, 1x100ml D40mm RB, 2/pk
13.8
-0.8
30314865
Rack, 1x 85ml D38mm RB, 2/pk
14.0
-0.6
30314866
Rack, 1x50ml D34mm RB, 2/pk
14.0
-0.6
30314867
Rack, 1x50ml D29mm RB, 2/pk
14.0
-0.6
30314868
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.5
30314869
Rack, 1x50ml D29mm Rim, 2/pk
13.9
-0.7
30314870
Rack, 1x30ml D25mm Rim, 2/pk
13.1
-1.5
30314871
Rack, 1x25ml D25mm RB, 2/pk
14.0
-0.6
30314872
Rack, 1x15ml D17mm FA, 2/pk
14.1
-0.5
30314873
Rack, 2x15ml D17mm FA, 2/pk
14.1
-0.5
30314874
Rack, 2x15ml D17mm Rim, 2/pk
14.0
-0.6
30314875
Rack, 4x15ml D17mm RB, 2/pk
14.0
-0.6
30314876
Rack, 7x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314877
Rack, 5x1.5ml D11mm, 2/pk
14.0
-0.6
30314878
Rack, 3x10ml D17mm Sar, 2/pk
14.0
-0.6
30314879
Rack, 4x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314880
Rack, 4x10ml D17mm Vac, 2/pk
14.0
-0.6
30314881
Bucket, 2x50ml D29mm FA, w / Rack,
2/pk
14.6
0.0
30314882
Bucket, 3x15ml D17mm FA, w / Rack,
2/pk
14.6
0.0
30314883
Cy to, Rectangular Bucket, 2/pk
6.3
-8.3
30314824
Rotor, Sw ing out,
2x 3MTP, ID
Blank
NO Bucket/Adapter
12.0
30314891
Adapter, 48x 1.5/2.0ml D11mm, 2/pk
12.0
0.0
30314825
Rotor, Angle,
6x 85ml RB, ID, Hi
Blank
NO Bucket/Adapter
10.3
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
9.6
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
9.6
-0.7
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.0
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
9.8
-0.5
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
9.3
-1.0
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
10.3
0.0
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
9.5
-0.8
30314826
Rotor, Angle,
6x 85ml RB, ID
Blank
NO Bucket/Adapter
11.5
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.9
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
10.4
-1.1
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.9
-0.6
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
10.6
-0.9
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
10.4
-1.1
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
11.1
-0.4
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
10.4
-1.1
30314827
Rotor, Angle,
Blank
NO Bucket/Adapter
9.2
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-43
4x 85ml RB, ID, Hi
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
8.5
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
8.4
-0.8
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
8.9
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
8.6
-0.6
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
8.3
-0.9
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
7.5
-1.7
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
8.3
-0.9
30314828
Rotor, Sw ing out,
16x 50ml, ID
Blank
NO Bucket/Adapter
16.5
30304375
Rack, 4x50ml D29mm FA, 2/pk
16.5
0.0
30314583
Rack, 10x 15ml D17mm FA, 2/pk
16.5
0.0
30314829
Rotor, Angle,
10x 50ml FA, ID
Blank
NO Bucket/Adapter
13.0
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
12.8
-0.2
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.4
-2.6
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
8.9
-4.1
30314830
Rotor, Angle,
6x 50ml RB/FA, ID
Blank
NO Bucket/Adapter
11.0
30130891
Adapter, 1x 30ml D26mm, 2/pk
10.7
-0.3
30130892
Adapter, 1x 16ml D18mm, 2/pk
10.3
-0.7
30130893
Adapter, 1x 15ml D17mm, 2/pk
10.6
-0.4
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.4
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
8.3
-2.7
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
6.7
-4.3
30314831
Rotor, Angle,
6x 50ml RB, ID, Hi
Blank
NO Bucket/Adapter
8.4
30130891
Adapter, 1x 30ml D26mm, 2/pk
8.2
-0.2
30130892
Adapter, 1x 16ml D18mm, 2/pk
7.9
-0.5
30130893
Adapter, 1x 15ml D17mm, 2/pk
8.0
-0.4
30314892
Adapter, 1x 15ml D17.5mm FA, 2/pk
7.7
-0.7
30314832
Rotor, Angle,
30x 15ml RB/FA,
ID
Blank
NO Bucket/Adapter
12.5
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.3
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.5
-2.0
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
9.0
-3.5
30314834
Rotor, Angle,
12x 15ml RB/FA,
ID
Blank
NO Bucket/Adapter
11.0
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.6
-0.4
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
9.1
-1.9
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
7.7
-3.4
30314835
Rotor, Angle,
44x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.7
-0.7
30314836
Rotor, Angle,
30x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
9.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
9.1
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
8.4
-1.0
30314838
Rotor, Angle,
24x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314839
Rotor, Angle,
12x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
6.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
6.3
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
5.6
-0.9
30314900
Adapter, 1x 1.5ml for ov er16000x g,
6/pk
6.4
-0.1
30553085
Rotor Sw ingOut
Blank
NO Bucket/Adapter
19.3
ES-44 Centrífuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4x 750ml D99mm
ID SL
30553138
Adapter 32x 7ml D13mm RBx 2 V1
18.7
-0.6
30553135
Adapter 18x 15ml D17mm RBx 2 V1
18.8
-0.5
30553132
Adapter 8x 30ml D26mm RBx 2 V1
19.1
-0.2
30553131
Adapter 7x 50ml D29mm RBx 2 V1
19.1
-0.2
30553127
Adapter 3x 100ml D41mm RBx 2 V1
19.1
-0.2
30553125
Adapter 1x 250ml D62mm FBx 2 V1
18.9
-0.4
30553126
Adapter 1x 225/190ml D61mm FAx 2
V1
19.1
-0.2
30553128
Adapter 5x 50ml D29mm FAx 2 V1
19.1
-0.2
30553130
Adapter 7x 50ml D29mm FAx 2 V1
19.1
-0.2
30553133
Adapter 14x 15ml D17mm FAx 2 V1
19.2
-0.1
30553134
Adapter 17x 15ml D17mm FAx 2 V1
19.0
-0.3
30553140
Adapter 18x 5/10ml D16mm
Vac/Sarx 2 V1
19.0
-0.3
30559377
Adapter 38x 1.1/1.4ml D8mm Sarx2
V1
18.9
-0.4
30553136
Adapter 21x 10ml D16mm RBx 2 V1
18.9
-0.4
30553139
Adapter 27x 1.8/7ml D13mm
Vac/Sarx 2 V1
18.8
-0.5
30553086
Rotor Angle
4x 500ml D69mm
ID
Blank
NO Bucket/Adapter
14.5
30564850
Adapter 1x 250ml D61.5mm FBx2 V1
13.7
-0.8
30559417
Adapter 1x 175/225ml D61.5mm
FAx 2 V1
13.4
-1.1
30559419
Adapter 1x 50ml D29.5mm FAx 2 V1
12.4
-2.1
30559416
Adapter 1x 250ml D60.5mm CBCx 2
V1
12.6
-1.9
30559421
Adapter 4x 30ml D25.5mm RBx 2 V1
13.8
-0.7
30559420
Adapter 3x 50ml D29mm RBx 2 V1
14.3
-0.2
30559422
Adapter 6x 15ml D17mm FAx 2 V1
13.8
-0.7
Centfuga Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-45
11.7 Tabla 7: Formulario de devolución / Certificado de descontaminación
Incluya este formulario con todas las devoluciones de equipo y ensambles.
La declaración completa acerca de la descontaminación es un prerrequisito para la asunción y posterior procesamiento de la devolución. Si no
se incluye la explicación correspondiente, llevamos a cabo la descontaminación con costos a su cargo.
Apellido: _____________________________________
Organización / compañía: _____________________________________
Calle: _____________________________________
CÓDIGO POSTAL: _____________ lugar:____________________
Teléfono: _____________ fax:____________________
Correo electrónico: _____________________________________
Pos. Crowd
Decontaminated
object
Serial number Description / Comment
1
2
3
4
¿Hay partes de las mencionadas anteriormente en contacto con las siguientes sustancias?
Soluciones, búferes, ácidos o álcalis acuosos que ponen en riesgo la salud:………. No
Agentes potencialmente infecciosos: ……………………………………………..……….. No
Reactivos y solventes orgánicos: …………………………………………………………... No
Sustancias radioactivas: ……………………………………….………. α.. β.. γ.. No
Proteínas que ponen en riesgo la salud: …………………………………………..………. No
ADN: …………………………………………………………………………………..……….. No
¿Estas sustancias han tocado el equipo o ensamble? …………………………………… No
¿Cuál, en caso afirmativo?:
________________________________________________________________
Descripción de las medidas para la descontaminación de las partes mencionadas:
Confirmo la descontaminación adecuada:
Compañía/Departamento ____________ Lugar y fecha: ______________________________
Firma de la persona autorizada: ________________________________________
Llene en letras de
molde mayúsculas
3ODQFURTXLVGX)&)&)&)&
Figure. 1
&URTXLVGHODFDUWHGX)&5
Figure. 2
&DUWHFURTXLVGX)&5)&5
Figure. 3
1 Couvercle de la centrifugeuse
2 Fenêtre du rotor
3 Affichage
4 Étiquette de fonction
5 Contacteur secteur
6 Connexion de l'alimentation
7 Déverrouillage d'urgence
7
4
1
3
6
2
4
1
5
6
7
2
3
4
1
5
6
7
2
3
Étiquette de fonction
Étiquette de fonction pour les FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figure. 4
Figure. 5
1. Affichage ACL
2. Modèle de configuration du
programme
3. Démarrage de la centrifugation
4. Sélection et modèle RPM/RCF
5. Sélection et modèle d'intensité
de l'accélération/de la
célération
6. Modèle de configuration de la
température
(FC5515R seulement)
7. Modèle de glage de l'heure
8. Informations de la configuration
du stockage
9. Centrifugation par rotation
courte/rapide
10. Déverrouillage du couvercle
11. Arrêt de la centrifugation/de la
configuration
12. Réglage du bouton/du cadran :
Changement du numéro
Affichage ACL
L'image suivante présente les éléments individuels de l'affichage ACL.
Figure. 6
1
2
3
4
5
6
7
8
9
10
11
12
Figure. 7
Affichage des champs:
A-1 Affichage champ « RPM/RCF »
A-2 Affichage champ « Acc/Déc » « Service »
A-3 Affichage champ « Heure/Prog »
A-4 Affichage champ « Temp »
Messages/logos de l'affichage des champs
M1 « fermer »
M2 « ouvrir »
M3 « rotor »
M4 « Rotor »
M5 « tr/min »
M6 « rcf »
M7 « accél »
M8 « cél »
M9 « rayon »
M10 « programme »
M11 « erreur »
M12 « service »
M13 « h m s »
M14 « température »
M15 « pré-refroidissement »
M14
M5
M6
M4
A-1
A-2
A-3
A-4
M2
M3
M1
M7
M8
M12
M15
M9
M11
M10
M13
Rotor Tableau
Affichage
du
numéro
de rotor
N° de
command
e
Capacité
Compatible
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
61
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
20
30314820
Rotor, Swing out, 4x250ml, ID
21
30314821
Rotor, Angle, 6x250ml FB, ID
22
30314822
Rotor, Swing out, 4x100ml, ID
23
30314823
Rotor, Swing out, 4x100ml, ID, Sealable
24
30314824
Rotor, Swing out, 2x3MTP, ID
25
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
26
30314826
Rotor, Angle, 6x85ml RB, ID
27
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
28
30314828
Rotor, Swing out, 16x50ml, ID
29
30314829
Rotor, Angle, 10x50ml FA, ID
30
30314830
Rotor, Angle, 6x50ml RB/FA, ID
31
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
32
30314832
Rotor, Angle, 30x15ml RB/FA, ID
33
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
34
30314834
Rotor, Angle, 12x15ml RB/FA, ID
35
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
36
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
38
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
39
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
40
30314840
Rotor, Angle, 64x0.5ml, ID
41
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
85
30553085
Rotor SwingOut 4x750ml D99mm ID SL
86
30553086
Rotor Angle 4x500ml D69mm ID
TABLE DES MATIÈRES
1. INTRODUCTION............................................................................................................................................................................ 1
1.1 DESCRIPTION ............................................................................................................................................................................... 1
1.2 CARACTÉRISTIQUES ........................................................................................................................................................................... 1
1.3 DEFINITION DES SYMBOLES ET DES INDICATEURS D'AVERTISSEMENTS ....................................................................................................... 1
1.4 PRECAUTIONS DE SECURITE ................................................................................................................................................................ 2
1.4.1 Utilisateur ........................................................................................................................................................................... 2
1.4.2 Rotor et accessoires ........................................................................................................................................................... 2
1.4.3 Mesures pour votre protection ......................................................................................................................................... 2
1.4.4 Exclusions d'influences environnementales suivantes ................................................................................................... 3
1.4.5 Mesures pour un fonctionnement sans danger .............................................................................................................. 3
1.4.6 Dangers et précautions ..................................................................................................................................................... 3
2. INSTALLATION.............................................................................................................................................................................. 4
2.1 DEBALLAGE...................................................................................................................................................................................... 4
2.1.1 Emballage de livraison ...................................................................................................................................................... 4
2.2 SÉLECTION DE L'EMPLACEMENT .......................................................................................................................................................... 4
2.3 INSTALLATION .................................................................................................................................................................................. 4
2.4 PRECAUTIONS DE SECURITE PENDANT LE FONCTIONNEMENT ET GARANTIE ................................................................................................ 5
3. FONCTIONNEMENT ..................................................................................................................................................................... 5
3.1 MONTAGE ET CHARGEMENT DU ROTOR ............................................................................................................................................... 5
3.1.1 Installation des rotors........................................................................................................................................................ 5
3.1.2 Chargement des rotors angulaires ................................................................................................................................... 6
3.1.3 Chargement des rotors horizontaux ................................................................................................................................ 7
3.1.4 Chargement et surcharge des rotors ............................................................................................................................... 7
3.1.5 montage du rotor .......................................................................................................................................................... 7
3.2 CONTROLE DU COUVERCLE................................................................................................................................................................. 7
3.2.1 Ouverture du couvercle ..................................................................................................................................................... 7
3.2.2 Verrouillage du couvercle ................................................................................................................................................. 8
3.3 PSÉLECTION.................................................................................................................................................................................. 8
3.3.2 Prélection des durées de centrifugation ...................................................................................................................... 9
3.3.3 Prélection de l'intensité de freinage et de l'accération........................................................................................... 9
3.3.4 Prélection de la temrature (Seulement les modèles réfrigés).......................................................................... 10
3.3.5 Pré-refroidissement (Seulement les moles réfrigérés)............................................................................................. 10
3.4 CORRECTION DU RAYON .................................................................................................................................................................. 11
3.5 PROGRAMMATION.......................................................................................................................................................................... 11
3.5.1 morisation des programmes ..................................................................................................................................... 11
3.5.2 Rappel des programmes mémorisés .............................................................................................................................. 12
3.5.3 Sortie du mode de programmation................................................................................................................................ 12
3.6 DEMARRAGE ET ARRET DE LA CENTRIFUGEUSE .................................................................................................................................... 12
3.6.1 marrage de la centrifugeuse ...................................................................................................................................... 12
3.6.2 La touche « STOP » .......................................................................................................................................................... 13
3.7 DETECTION D'UN DESEQUILIBRAGE.................................................................................................................................................... 13
4. PARAMÉTRAGE .......................................................................................................................................................................... 14
4.1 RÉGLAGES DE BASE ......................................................................................................................................................................... 14
4.1.1 Acs au mode « Données de fonctionnement ».......................................................................................................... 14
4.1.2 Indication de la température .......................................................................................................................................... 14
4.1.3 Signal Marche/Art ........................................................................................................................................................ 15
4.1.4 Volume de présélection du signal sonore...................................................................................................................... 15
4.1.5 lection de la sonorité du signal sonore de fin de centrifugation............................................................................. 16
4.1.6 Marche/Arrêt du son du clavier ..................................................................................................................................... 16
4.1.7 Appel des données de fonctionnement.......................................................................................................................... 16
5. MAINTENANCE........................................................................................................................................................................... 17
5.1 MAINTENANCE ET NETTOYAGE ......................................................................................................................................................... 17
5.1.1 Entretienral : ........................................................................................................................................................... 17
5.1.2 Nettoyage etsinfection de l'unité.............................................................................................................................. 18
5.1.3 Nettoyage etsinfection du rotor................................................................................................................................ 18
5.1.4 sinfection des rotors en aluminium ........................................................................................................................... 18
5.1.5 sinfection des rotors PP .............................................................................................................................................. 18
5.1.6 Bris de verre ...................................................................................................................................................................... 19
6. PANNAGE ............................................................................................................................................................................... 19
6.1 MESSAGE D'ERREUR : CAUSES/SOLUTIONS......................................................................................................................................... 19
6.2 ÉTUDE DES MESSAGES D'ERREUR POSSIBLES ET LEURS SOLUTIONS .......................................................................................................... 19
6.2.1 Liration du couvercle pendant une panne d'alimentation (libération en cas d'urgence) .................................... 19
6.2.2 Description du système de messages d'erreur.............................................................................................................. 20
7. RÉCEPTION DES CENTRIFUGEUSES POUR RÉPARATION........................................................................................................ 21
8. TRANSPORT, STOCKAGE ET MISE AU REBUT .......................................................................................................................... 21
8.1 TRANSPORT ................................................................................................................................................................................... 21
8.2 STOCKAGE ..................................................................................................................................................................................... 21
9. DONNEES TECHNIQUES............................................................................................................................................................. 23
9.1 SPECIFICATIONS.............................................................................................................................................................................. 23
9.1.1 Centrifugeuse FC5714 ...................................................................................................................................................... 23
9.1.2 Centrifugeuse FC5718 ...................................................................................................................................................... 24
9.1.3 Centrifugeuse FC5718R.................................................................................................................................................... 25
9.1.4 Centrifugeuse FC5816 ...................................................................................................................................................... 26
9.1.5 Centrifugeuse FC5816R.................................................................................................................................................... 27
9.1.6 Centrifuge FC5916 ............................................................................................................................................................ 28
9.1.7 Centrifuge FC5916R.......................................................................................................................................................... 29
9.2 SCHEMAS ET DIMENSIONS................................................................................................................................................................ 30
10. CONFORMI ........................................................................................................................................................................ 31
11. ANNEXE.................................................................................................................................................................................. 32
11.1 TABLE 1: POIDS NET AUTORISE ................................................................................................................................................ 33
11.2 TABLE 2: TEMPERATURES LES PLUS FAIBLES A LA VITESSE MAXIMUM .......................................................................... 35
11.3 TABLE 3: VITESSE ET VALEURS RCF MAXI POUR LES ROTORS AUTORISES ...................................................................... 35
11.4 TABLE 4: DUREES D'ACCELERATION ET DE DECELERATION .................................................................................................... 39
11.5 TABLEAU 5 : MESSAGES D'ERREUR ........................................................................................................................................... 41
11.6 TABLE 6: CORRECTION DU RAYON ........................................................................................................................................... 42
11.7 TABLEAU 7 : FORMULAIRE DE RACHAT/CERTIFICAT DE DECONTAMINATION .......................................................................................... 46
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-1
1. INTRODUCTION
1.1 DESCRIPTION
Merci d'avoir choisi ce produit OHAUS.
Tous les symboles indiquent des instructions de sécurité et des emplacements présentant des situations
potentiellement dangereuses. Veuillez lire le manuel complètement avant d'utiliser les centrifugeuses Frontier
TM
MultiPro pour éviter un fonctionnement incorrect.
Les centrifugeuses Frontier
TM
MultiPro ont été conçues pour la séparation de matériaux ou de mélanges de
différentes densités.
Les centrifugeuses OHAUS sont conçues pour être utilisées exclusivement en intérieur et par du personnel qualif.
1.2 Caractéristiques
Les modèles FC5714/FC5718/FC5816/FC5916 sont des centrifugeuses universelles non-réfrigées.
Les modèles FC5718R/FC5816R/FC5916R sont des centrifugeuses universelles frigées.
Les trois modèles sont proposés avec deux tensions d’alimentation, 230 V ou 120 V
Les centrifugeuses sont utilisées avec des rotors horizontaux ou angulaires
Les réglages sont accessibles via des boutons et lections via le régleur central. Toutes les valeurs préréglées
et les valeurs courantes apparaissent en continu sur l’écran ACL.
La centrifugeuse est entraînée par un moteur asynchrone sans entretien
Les dones techniques taillés sont consultables dans «Données techniques»
1.3 finition des symboles et des indicateurs d'avertissements
Les remarques de sécurité sont signalées par des mots indicateurs et par des symboles d'avertissement. Elles
présentent les problèmes et les avertissements de sécurité. L'inobservation des remarques de sécurité peut
conduire à des accidents, à l'endommagement de l'instrument, à des dysfonctionnements et à des sultats
erronés.
Le niveau du danger fait partie d'une note de sécurité et alise la distinction entre les sultats possibles d'une
inobservation pour chacun d'entre eux.
Mots indicateurs
DANGER
Possibilité d'un accident grave, voire mortel, s'il n'est pas évité.
AVERTISSEMENT
Pour une situation dangereuse avec un risque moyen pouvant être à l'origine
d'un accident ou d'un décès, si elle n'est pas évitée.
ATTENTION
Pour une situation dangereuse avec un risque faible, pouvant être à l'origine
d'un endommagement de l'appareil ou des biens, de perte des dones, ou
de blessures si elle n'est pas évitée.
ATTENTION
Pour des informations importantes sur le produit. Peut conduire à des
tériorations de l'équipement si elle n'est pas évitée
NOTE
Pour des informations utiles sur le produit
Symboles d'avertissement
Danger général Danger de choc électrique
Tension alternative Risques biologiques
Explosion Écrasement
FR-2 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Signalisation informative et d'avertissement sur la surface de la centrifugeuse
Quatre supports doivent être utilisés à tout moment sur les quatre
emplacements de rotation des rotors, sinon des tériorations de la
centrifugeuse se produiront. De telles détériorations ne seront pas couvertes
par la garantie de ce produit.
Attention ! Vérifier le serrage de l'écrou du rotor avant chaque utilisation.
Retirer le bouchon secteur avant d'ouvrir
l'enceinte ou d'effectuer un déverrouillage d'urgence.
Sens de la rotation, rotation horaire pour l’entraînement du rotor
Référence pour le chargement des rotors
1.4 Précautions de sécurité
1.4.1 Utilisateur
Les centrifugeuses Ohaus sont exclusivement desties à une utilisation intérieure et par un personnel qualifié.
Cet appareil ne peut être utilisé que par un personnel ment formé. Ce personnel devra lire intégralement le
manuel de fonctionnement et se familiariser au fonctionnement de l'appareil.
1.4.2 Rotor et accessoires
Seuls des rotors et des accessoires d'origine d'OHAUS seront utilisés. Toute autre utilisation ou utilisation
envisagée est considérée comme inadéquate. OHAUS n'est pas responsable des détériorations sultant d'une
utilisation inaquate.
ATTENTION :
Lire tous les avertissements de sécurité avant d’effectuer l’installation, les connexions ou la maintenance
de cet équipement. Si ces avertissements ne sont pas respectés, des accidents et/ou des dommages
aux biens peuvent en résulter. Conserver ces instructions pour référence ultérieure.
1.4.3 Mesures pour votre protection
AVERTISSEMENT : Ne jamais travailler dans un environnement pouvant présenter des dangers
d'explosion ! L'enceinte de l'instrument n'est pas étanche aux gaz. (Danger d'explosion en présence
d'étincelles, corrosion provoqe par l'introduction de gaz)
AVERTISSEMENT : Lors de l'utilisation de produits chimiques et de solvants, se conformer aux
instructions du fabricant et aux gles générales de sécurité du laboratoire.
AVERTISSEMENT :La centrifugeuse n'est pas étanche. Des mesures de protection adaptées seront
prises lors de l'utilisation de la centrifugeuse pour des échantillons infectieux ou pathogènes. Des
précautions de sécurité appropriées doivent être appliquées lors de la manipulation de ces échantillons.
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-3
1.4.4 Exclusions d'influences environnementales suivantes
Vibrations puissantes
Lumière directe du soleil
Humidité atmosphérique surieure à 80 %
Présence de gaz corrosifs
Températures supérieures à 35 °C et inférieures à 2 °C
Champs électriques ou magtiques puissants :
AVERTISSEMENT :
Risque de chocs électriques dans le btier. Le btier doit être ouvert uniquement par un personnel
qualifet autorisé. Enlever toutes les connexions d’alimentation à l’unité avant l'ouverture.
1.4.5 Mesures pour un fonctionnement sans danger
Ne jamais dévisser les deux moitiés de l'enceinte
Sécher immédiatement tout déversement de liquide. L'instrument n'est pas étanche
Vérifier que la plage de tension d'entrée et le type de prise de l'équipement sont compatibles avec la tension
secteur locale.
Brancher le câble d'alimentation uniquement à une prise d'alimentation électrique correctement mise à la masse.
Utiliser uniquement un câble d'alimentation avec une puissance nominale surieure aux scifications sur
l'étiquette de l'équipement.
Positionner l'équipement pour qu'il soit facile de connecter le câble d'alimentation de la prise d'alimentation
électrique.
S'assurer que le cordon d'alimentation n’est pas un obstacle potentiel ou ne présente aucun danger de chute.
Cet équipement est exclusivement desti à un usage en intérieur. Utiliser l'équipement seulement dans des
lieux secs.
Utiliser uniquement des accessoires approuvés.
Faire fonctionner l'équipement uniquement dans les conditions ambiantes spécifiées dans ces instructions.
Déconnecter l'équipement de l'alimentation électrique pour le nettoyer.
Ne pas utiliser l'équipement dans des environnements dangereux ou instables.
La maintenance doit être exclusivement exécutée par un personnel agréé.
1.4.6 Dangers et pcautions
Les pcautions suivantes doivent être observées afin de protéger les personnes et l'environnement :
Pendant la centrifugation et la mise en œuvre de matières dangereuses, la présence de personnes est interdite
dans un rayon de 30 cm autour de la centrifugeuse conformément aux réglementations EN 61010-2-020.
Les FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ne sont pas protégées contre les
explosions et ne doivent pas être utilisées dans des lieux ou dans des zones susceptibles aux explosions. La
centrifugation de substances inflammables, explosives ou radioactives qui réagissent chimiquement en
produisant une énergie élevée est strictement interdite. La cision finale quant aux risques associés à
l'utilisation de telles substances incombe à l'utilisateur de la centrifugeuse.
Ne pas centrifuger de matières toxiques ou pathogènes sans précautions de sécurité adéquates, c'est-à-dire la
centrifugation de godets/de tubes présentant une étancité peu hermétique ou défectueuse, ce qui est
strictement interdit. L'utilisateur a l'obligation d'appliquer des procédures appropriées de sinfection si des
substances dangereuses ont contaminé la centrifugeuse et/ou ses accessoires. Lors de la centrifugation de
substances infectieuses, l'utilisateur prêtera attention aux précautions nérales de laboratoire. Si nécessaire,
contacter le responsable de la sécurité !
Il est interdit de faire fonctionner la centrifugeuse avec d'autres rotors que ceux répertoriés pour cette unité.
Le couvercle de la centrifugeuse ne doit jamais être ouvert tant que le rotor est en fonctionnement ou en rotation à
une vitesse surieure à 2 m/s
FR-4 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
1.4.7 Abréviations utilisées dans ce manuel
2. INSTALLATION
2.1 Déballage
Sortir avec pcaution la centrifugeuse et chacun de ses composants hors de l'emballage. Les composants inclus
varient en fonction du modèle de centrifugeuse (voir le tableau ci-dessous). Conserver l'emballage afin d'assurer
un stockage et un transport en toute sécurité. Le manuel d'instructions doit toujours se trouver à proximité de la
centrifugeuse !
Les rotors/les accessoires seront emballés séparément.
AVERTISSEMENT : Risques associés au levage. Risque de blessures si une seule
personne soulève l'équipement. Se faire aider lorsque l'équipement est soule ou
placé.
On préconise qu’il y ait au moins quatre personnes, voire davantage, pour déplacer et installer ces
trois unis ou d’utiliser un matériel de transport adapté. Consulter la section 8.3 pour davantage de
précisions sur la techniques de lavage pour la sortie de lemballage.
2.1.1 Emballage de livraison
2.2 Sélection de l'emplacement
REMARQUE !
Éviter les vibrations excessives, les sources de chaleur, les courants d'air ou les variations rapides de
température.
La centrifugeuse doit être installée sur une surface plane, de niveau et résistante, si possible dans une
armoire/sur une table de laboratoire ou sur une surface solide sans vibrations.
Pendant la centrifugation, la centrifugeuse doit être placée de manière à présenter un espace minimum de 30
cm de chaque côté de l'unité conformément aux normes EN 61010-2-020.
La centrifugeuse ne doit pas être positionnée à proximité d'une fenêtre ou d'un chauffage, elle pourrait être
exposée à une chaleur excessive, en effet les performances de l'unité se fondent sur une température ambiante
de 23 °C.
2.3 Installation
Suivre ces étapes :
Vérifier que l'alimentation électrique correspond à celle spécifiée sur l'étiquette signalétique du fabricant, située
sur le panneau arrière.
Quantité
Description
1
Centrifugeuse
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
1
Câble d’alimentation
1
Carte de garantie
1
Manuel d’instructions / Guide de démarrage rapide
1
Cdu rotor
Symbol/Abbreviations Unit Description
RPM
[min
-1
]
rpm
revolutions per minute
RCF [x g]
relative centrifugal force
PCR
PCR Polymerase chain reaction
PP - Polypropylen
PC - Polycarbonat
accel - acceleration
decel - deceleration
prog - program
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-5
Pour les modèles FC5714/FC5718/FC5816, la ligne d'alimentation doit être protégée par un disjoncteur 10 A
cote de circuit (type K).
Pour les modèles FC5718R/FC5816R /FC5916/FC5916R, la ligne d'alimentation doit être protégée par un
disjoncteur 16 A cote de circuit (type K)
En cas d'urgence, un commutateur de sûredoit être installé en dehors de la pièce afin de déconnecter
l'alimentation sur l'unité.
La centrifugeuse doit être connectée à une prise secteur reliée à la terre.
Raccorder la centrifugeuse au seau.
(La prise destie au câble d’alimentation doit être d’un accès facile, de même que facile à brancher.
Mettre l'instrument sous tension avec le commutateur d'alimentation.
Ouvrir le couvercle en utilisant le bouton Door Open (porte ouverte).
Retirer le dispositif sécurisant le mécanisme d'entraînement du moteur.
2.4 Précautions de sécurité pendant le fonctionnement et garantie
Ne pas faire fonctionner la centrifugeuse si elle n'est pas correctement installée.
Ne pas s'appuyer contre la centrifugeuse pendant le fonctionnement.
Ne pas rester dans la zone de sécurité de 30 cm autour de l'équipement plus longtemps que nécessaire pour
des raisons de fonctionnement.
Ne laisser aucun produit potentiellement dangereux dans la zone de sécurité de 30 cm autour de l'équipement.
Ne pas faire fonctionner la centrifugeuse lorsqu'elle est montée (sans son enceinte).
Ne pas faire fonctionner la centrifugeuse lorsque des composants mécaniques ou électriques ont subi des
altérations.
Ne pas utiliser d'accessoires tels que des rotors ou des godets qui ne sont pas exclusivement approuvés par
OHAUS Corporation, hormis les tubes pour centrifugeuse disponibles commercialement et fabriqués en verre
ou en plastique.
Ne pas centrifuger de substances fortement corrosives qui pourraient endommager ou fragiliser les matériels.
Ne pas faire fonctionner la centrifugeuse avec des rotors ou des godets présentant des signes de corrosion ou
de détériorations mécaniques.
Le fabricant est responsable de la sécurité et de la fiabilité de la centrifugeuse, seulement si :
L'unité est exploitée conformément à ce manuel d'instructions.
Des modifications, des parations ou d'autres réglages sont alisés par un personnel agréé et l'installation
électrique est conforme à la réglementation en la matière.
REMARQUE!
Garantie
La centrifugeuse a fait l’objet d’une vérification complète et de contrôles de qualité. Dans l’éventualité, peu
probable, de vices de fabrication, la centrifugeuse et les rotors sont couverts par la garantie. Ladite garantie est
nulle en cas de mauvaise utilisation, de dommages et de négligence, également dans le cas de l’utilisation de
pièces de rechange et d’accessoires inadaptés, voire de la modification non-autorisée de la machine.
Le droit d’apporter des modifications relève de la seule responsabilité du fabricant en vue d’apporter des
améliorations!
3. FONCTIONNEMENT
3.1 Montage et chargement du rotor
3.1.1 Installation des rotors
Nettoyer l'arbre d'entraînement ainsi que le collier de serrage avec un chiffon propre et sans graisse. Placer le rotor
sur l'arbre d'entraînement. (Voir figure ci-dessous). Ne pas oublier que le rotor est intégralement installé sur l'arbre
du moteur.
FR-6 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figure. 8
Figure. 9 Figure. 10
Maintenir le rotor d'une main et fixer celui-ci sur l'arbre en tournant la vis de fixation dans le sens horaire. Serrer
l'écrou de fixation avec la cpour rotor fournie (voir les figures 9 et 10).
Nous fournirons un outil pour les rotors sans écrou avec la centrifugeuse, l'outil pour le rotor avec écrou sera fourni
en même temps que le rotor.
ATTENTION :
Vérifier que la vis de fixation est correctement installée avant chaque centrifugation. (Se reporter à la
figure 11)
Ne pas faire fonctionner la centrifugeuse avec des rotors ou
des godets présentant des signes de corrosion ou de
détériorations mécaniques.
Ne pas faire fonctionner la centrifugeuse avec des substances
fortement corrosives, ce qui pourrait endommager le rotor, les
godets et les matériels.
En cas de questions, contacter le fabricant !
3.1.2 Chargement des rotors angulaires
Les rotors doivent être chargés symétriquement et avec un poids égal (voir la figure ci-dessous). L'adaptateur ne
peut être charqu'avec les cipients appropriés. La différence de poids entre les cipients remplis doit être
autant que faire se peut réduite au minimum. C'est pourquoi nous recommandons leur pesée avec une balance.
Ceci réduit l'usure de l'entrnement et le bruit de centrifugation.
La charge maximum par orifice est définie sur chaque rotor.
Figure. 11 INCORRECT (6 tubes) Figure. 12 CORRECT
Écrou du rotor
Outil du rotor
avec écrou
Outil du rotor
sans écrou
Couvercle à
enclenchement
Couvercle à
vissage
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-7
3.1.3 Chargement des rotors horizontaux
Le chargement des godets ou des cuves doit être réalisé conformément à la figure ci-dessous. On peut exploiter, par exemple,
un rotor à quatre logements avec seulement deux godets. Néanmoins, les godets chargés doivent être en vis-à-vis. S’assurer
que les godets non chargés soient également placés à lintérieur du rotor (cf. ci-dessous).En principe, il n’est pas permis de
mettre en œuvre les rotors jusqu’à ce que tous les godets ou râteliers soient installés dans le rotor. Les boulons de rotor doivent
être lubrifiés avec du «High TEF Oil». Les tubes à essai doivent être remplis visuellement de fon uniforme et disposés dans
les orifices ou les râteliers à tubes. La différence en poids des godets chargés de doit pas dépasser environ 1,0 g.
ATTENTION!
Il n’est permis de mettre en œuvre les rotors horizontaux uniquement que si tous les emplacements sont
occupés, soit avec quatre godets, soit avec quatre porteurs; ne pas mélanger godets et porteurs !!.
ATTENTION!
Ne pas mettre en œuvre la centrifugeuse avec des rotors ou des godets présentant des signes de corrosion
ou de fatigue mécanique.
Ne pas mettre en œuvre avec des substances extrêmement corrosives qui pourraient endommager le rotor
ou des godets. Contacter le fabricant pour toute question!
Figure.13 WRONG Figure.14 CORRECT
3.1.4 Chargement et surcharge des rotors
Tous les rotors approuvés sont répertoriés avec leur vitesse maximum ainsi que leur poids de remplissage
maximum sur le « tableau 2 du poids net autorisé » (voir ANNEXE).
La charge maximum autorisée pour un rotor, détermie par le fabricant, ainsi que sa vitesse maximum autorisée
(consulter l'étiquette sur le rotor), ne doivent pas être dépassées. Les liquides chargés sur les rotors doivent
présenter une densité homone maximum de 1,2 g/ml ou moins lorsque le rotor fonctionne à sa vitesse maximum.
Pour centrifuger des liquides d'une densité plus élevée, la vitesse doit être réduite conformément à la formule
suivante :
Vitesse duite n
rouge
=
1,2
𝑑𝑒𝑛𝑠𝑖𝑡é 𝑝𝑙𝑢𝑠 é𝑙𝑒𝑣é𝑒
x vitesse maxi (n
maxi
) du rotor
Exemple :
n
rrouge
=
1,2
1,7
x 4 000 = 3 360 tr/min
En cas de questions, contacter le fabricant SVP !
Afin de déterminer la force centrifuge relative (FCR/Force g) pour un adaptateur particulier, on peut calculer en DIN
58 970 à l’aide de la formule jointe:
FCR = 1.117862 *10
5
*n
2
*r
max
n: tours par minute (tr/min.)
r
max
: rayon max. de centrifugation en cm à la base des tubes.
3.1.5 Démontage du rotor
Desserrer complètement l'écrou de fixation du rotor (dévisser au-dedu point de sistance) et soulever le rotor
verticalement hors de la centrifugeuse. (Voir les figures 10 et 11)
3.2 Contrôle du couvercle
3.2.1 Ouverture du couvercle
Après une centrifugation, lors de la fermeture du couvercle de la centrifugeuse, le mot "close" (M1) appart sur
l'affichage "RPM | RCF" (A-1). De plus, si un rotor se trouve dans la centrifugeuse, le mot "rotor"(M3) apparaît
FR-8 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
ainsi que le numéro de code du rotor s'y trouvant installé "71" (M4). En l'absence de rotor dans la centrifugeuse, le
mot "rotor"(M3) clignote et appart en suppment le mot "no" (M4). En appuyant sur la touche "Door Open" (7),
l'utilisateur pourra libérer le couvercle de la centrifugeuse. Aussitôt que le couvercle électromagnétique est
complètement libéré, le mot "open" (M2) appart. Il est donavant possible d'ouvrir le couvercle de la
centrifugeuse.
Se reporter à la figure 15 ci-dessous pour férence.
Figure. 15
Pendant la centrifugation, l'utilisateur peut rechercher le type de rotor à tout moment en appuyant sur la touche
"Door Open" (7).
3.2.2 Verrouillage du couvercle
Le couvercle ne doit être que ts légèrement abaissé. Un verrouillage électromagnétique du couvercle le referme
et simultanément, le mot "open" (M2) disparaît (voir la figure 15).
Afin de signaler que la centrifugeuse est sur le point de marrer, le mot "close" (M1) appart sur l'affichage
"RPM | RCF" (A-1). Simultanément le mot "rotor" (M3) s'affiche ainsi que le de code du rotor installé dans la
centrifugeuse, "no 71" (M4). Ainsi, toutes les dones scifiques du rotor, telles que la vitesse maximum,
l'accélération, etc., sont adoptées.
ATTENTION :
Ne pas se coincer les doigts entre le couvercle et l'appareil ou le mécanisme de verrouillage lors de la
fermeture du couvercle !
3.3 Prélection
3.3.1 Psélection de la vitesse/de la valeur RCF
Cette présélection est activée au moyen de la touche "RPM | RCF" (4) (voir la figure 16 ci-dessous). En appuyant
sur la touche une fois, le mot "rpm" (M5) clignote. En appuyant sur la touche deux fois, il est possible de
présélectionner les forces centrifuges. Le mot clignotant "rcf" (M6) appart alors. Il est possible de finir les
valeurs désies avec le bouton de réglage (1). Sur l'affichage (A-1), la valeur gulée est présentée en
permanence, avant, pendant et après la centrifugation.
Figure. 16
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-9
Tant qu'aucun rotor n'est inséré, la vitesse est glable entre 200 tr/min et le nombre maximum de révolutions de la
centrifugeuse.
Si un rotor se trouve dans la centrifugeuse, la vitesse ne peut être présélectione que jusqu'au maximum autorisé
de volutions de ce rotor. C'est la même que la présélection de la valeur RCF. La plage de réglage se situe entre
la force centrifuge à 200 tr / min et la force centrifuge maximale admissible du rotor.
Voir le « Tableau 3 : vitesse maximum et valeurs RCF des rotors utilisés » (voir ANNEXE). Toutes les valeurs
importantes sont
répertoriées ici.
ATTENTION :
L'utilisateur doit vérifier la rotation maximum autorisée des tubes à essai avec le fabricant.
3.3.2 Psélection des durées de centrifugation
La durée de centrifugation peut être présélectione selon trois différentes plages de 10 secondes à 99 heures 59
minutes.
1. Plage de 10 secondes jusqu'à 59 minutes 50 secondes par étapes de 10 secondes
2. Plage d'une heure jusqu'à 99 heures 59 minutes par étapes d'une minute
3. Le fonctionnement en continu "cont" peut être interrompu avec la touche "Stop" (10) (voir la figure 17).
La durée de fonctionnement peut être présélectione avec le couvercle ouvert ou fermé.
Pour activer le glage de la durée de fonctionnement, appuyez sur la touche "Time" (6).
Sur l'affichage « Time/Prog" (A-3), l'indication "m : s" ou "h : m" clignote en fonction des glages précédents.
Pour définir la valeur désie, utiliser le bouton de glage (1). Après avoir dépassé 59 min 50 s, l'indication passe
automatiquement sur "h : m". Après avoir dépassé 99 heures 59 min, le mot "cont" appart sur
l'affichage "Time/Prog" (A-3). Ce fonctionnement en continu ne peut être interrompu qu'en appuyant sur la touche
"Stop" (10). Le décompte temporel commence aussitôt que la vitesse prescrite est atteinte.
L'affichage présente toujours la durée de fonctionnement restante. (Voir la figure 17)
Figure. 17
3.3.3 Psélection de l'intensité de freinage et de l'accélération
Cette fonction est activée avec la touche "Accel/Decel" (5) (voir la figure 18).
En appuyant une fois sur la touche, le mot "accel" (M7) clignote sur l'affichage "Acc/Dec" (A-2). L'accélération
sirée peut être présélectionnée en réglant le bouton (1). La valeur 0 est équivalente à la plus faible accélération
et la valeur 9 correspond à la plus élevée.
En appuyant deux fois sur la touche "Accel/Decel" (5), le mot "decel"(M8) appart sur l'affichage "Acc/Dec" (A-
2). Dorénavant, l'intensité de freinage sirée peut être présélectionnée en réglant le bouton (1). La valeur 9 est
équivalente à la durée de freinage la plus courte et la valeur 0 correspond à la plus longue.
Voir le « Tableau 5 : Durées d'aclération et de cération » (ANNEXE). Les durées d'aclération et de
célération pour les étapes 0 à 9 d'accélération et de célération concernant les rotors utilisés sont présentées
ici.
FR-10 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figure. 18
3.3.4 Psélection de la température (Seulement les modèles frirés)
Cette fonction est activée par la touche "Temp/Setup" (13). Après avoir appuyé sur cette touche de l'affichage
"Time/Prog", l'indication C" (A-4) clignote. Avec le bouton de réglage (1), la température de test sie peut
être présélectione par incments de 1 °C dans la plage de -20 °C à +40 °C.
La valeur est indiquée en permanence sur l'affichage (figure 19), avant, pendant et après la centrifugation.
Prendre note des températures les plus basses des rotors à la vitesse maximum !
Figure. 19
3.3.5 P-refroidissement (Seulement les modèles frirés)
Si les échantillons sont sensibles à la température, il est utile de pré-refroidir la centrifugeuse, le rotor et finalement
les godets à la température requise de fonctionnement. Pour ce faire, insérer le rotor retenu et prédéfinir la
température correspondante. En appuyant simultanément sur les touches "Temp/Setup" (13) (voir la figure 20) et
"Time" (6), l'utilisateur marrera une centrifugation. Pendant le fonctionnement, l'unité choisit automatiquement
une vitesse de rotation équivalente à 20 % de celle autorisée du rotor inséré. Après avoir atteint la température
prédéfinie, il est possible de quitter le pré-refroidissement avec la touche "Stop" (10).
En fonction du rotor inséré, le pré-refroidissement dure environ de 10 à 20 minutes.
Figure. 20
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-11
3.4 Correction du rayon
L'utilisation d'adaptateurs ou de réducteurs peut modifier le rayon de centrifugation de chaque rotor. Dans ce cas, il
est possible de modifier manuellement le rayon. Poursuivre de la manière suivante :
Fermer le couvercle et maintenir ensuite appuyées les touches "Time" (6) (voir la figure 21) et "Prog/Setup" (11).
Sur l'affichage "Time/Prog" (A-3), le mot "radius" (M9) apparaît. Avec le bouton de glage (1), l'utilisateur peut
présélectionner des corrections du rayon (voir Tableau 7, ANNEXE) par incrément de 0,1 centimètre. Après avoir
fini une correction du rayon, le mot "radius" (M9) appart. Ce mot sera visible jusqu'à ce que la correction du
rayon soit revenue à 0.
Figure. 21
3.5 Programmation
3.5.1 Mémorisation des programmes
Il est possible de mémoriser jusqu'à 99 centrifugations incluant tous les paramètres pertinents, notamment les
rotors utilisés. Il est possible d'utiliser un numéro de programme libre et de le rappeler à nouveau.
Installer le rotor nécessaire dans la centrifugeuse. En appuyant sur la touche "Prog/Setup" (11) de l'affichage
"Time/Prog" (A-3), le mot "programm" appart. Avec le bouton de glage (1), choisir le numéro de programme
retenu.
Si un numéro de programme est occupé, les mots "rotor" (M3) et "xx" (M4) appartront sur l'affichage
"RPM | RCF" (A-1). Pour les numéros de programme libres, 0 appart.
Figure. 22
Fermer le couvercle de la centrifugeuse. Procéder dorénavant selon la description précédente afin de finir tous
les paramètres importants de la centrifugation. Si le couvercle n'est pas ferlors de la mémorisation du
programme, les mots "FirSt" et "CLOSE Lid" (voir la figure 23) clignotent alternativement sur l'affichage
"RPM/RCF" (A-1). Si la centrifugation doit être démare sans mémorisation du programme, les mots "First" et
"PrESS StoreE" (voir la figure 24) clignotent alternativement sur l'affichage "RPM/RCF" (A-1).
FR-12 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figure. 23 Figure. 24
Pour adapter les données, appuyer sur la touche "Store" (12) (voir les figures 23 et 24) pendant 1 seconde
environ. Si le programme est correctement mémorisé, le mot "StorE" apparaît sur l'affichage "RPM/RCF" (A-1).
Ceci aura pour effet de faire dispartre le mot "programm" (M10).
Aussitôt aps avoir libé la touche "Store" (12), le mot "programm xx" (M10) réapparaît avec xx correspondant à
l'emplacement du programme choisi.
Si tous les numéros de programme sont occupés, l'utilisateur peut choisir un ancien numéro qui n'est plus
actuellement utile et y introduire simplement les nouveaux paramètres.
3.5.2 Rappel des programmes mémorisés
Pour rappeler des programmes mémorisés, appuyer sur la touche "Prog/Setup" (11) (voir la figure 25) alors que le
couvercle est fermé. Sur l'affichage "Time/Prog" (A-3), le mot "programm –"(M10) appart. Le numéro de
programme siré peut être présélection avec le bouton de glage (1).
Sur les affichages respectifs, les valeurs mémorisées de ce programme appartront.
Si le rotor dans la centrifugeuse n'est pas celui correspondant au programme présélection, le mot "rotor" (M3)
clignotera sur l'affichage "RPM | RCF" (A-1). Simultanément, le mot "FALSE" et le numéro du rotor mémorisé "
xx" (M4) clignoteront alternativement.
Figure. 25
3.5.3 Sortie du mode de programmation
Pour quitter le mode de programmation, appuyer simplement sur la touche "Prog/Setup" (11) (voir la figure 25).
Sur l'affichage "Time/Prog », le mot "programm" appartra.
Régler l'affichage sur "programm--" (M10) avec le bouton de glage (1).
3.6 Démarrage et arrêt de la centrifugeuse
3.6.1 marrage de la centrifugeuse
La centrifugeuse peut être démarrée avec la touche "Start" (9) (voir la figure 26) ou avec la touche "Quick Spin"
(8).Avec la touche "Start" (9), l'utilisateur peut démarrer des centrifugations mémorisées ou exécuter une
centrifugation avec des paramètres psélections manuellement.
Lorsque les durées de fonctionnement psélectionnées sont termies, la centrifugeuse s'artera
automatiquement.
Avec la touche "Quick Spin" (8), l'utilisateur peut démarrer des centrifugations qui ne dureront que quelques
secondes.
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-13
En appuyant sur la touche "Quick Spin" (8), la centrifugeuse accélère jusqu'à la rotation présélectionnée.
Sur l'affichage "Time/Prog" (A-3), la due de fonctionnement transmise est indiquée aussitôt que la touche
"Quick Spin" (8) est enfoncée.
En relâchant la touche "Quick Spin" (8), la centrifugeuse s'arte et la durée de fonctionnement est indiquée
jusqu'à ce que le couvercle soit ouvert.
Figure. 26
3.6.2 La touche « STOP »
Avec la touche "Stop" (10) (voir la figure 27), il est possible d'interrompre une centrifugation à tout moment. Après
avoir appuyé sur la touche, la centrifugeuse décélère conformément à l'intensité présélectione jusqu'à son arrêt
complet.
Figure. 27
3.7 Détection d'un déséquilibrage
Si la charge n'est pas également partie sur le rotor, l'entraînement s'interrompra pendant l'accélération. Le rotor
célère jusqu'à l'art complet.
Lorsque sur l'affichage « Time/Prog" (A-3), le mot "error” (M11) apparaît conjointement au numéro "01, la
différence de poids des échantillons est trop importante. Répartir les poids uniformément.
Charger le rotor selon la description des chapitres 3.1.2 et 3.1.3.
Lorsque sur l'affichage « Time/Prog" (A-3), le mot "error” appart conjointement au numéro "02" (voir la figure
28), ceci peut être à la raison suivante : Le contacteur de séquilibrage est défectueux.
Figure. 28
FR-14 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4. PARAMÉTRAGE
4.1 Réglages de base
4.1.1 Accès au mode « Données de fonctionnement »
Lors de l'utilisation de la centrifugeuse, les paramètres suivants doivent être définis :
- Indication de la température en °C ou °F
- Arrêt/Marche du signal acoustique
- Arrêt/Marche du son du clavier
- Psélection du volume sonore du signal
- Sélection du signal sonore modulé "end of run" (centrifugation terminée)
La centrifugeuse étant hors tension, appuyer simultanément sur les touches "Time"(6) et "Door Open" (7), et
mettre la centrifugeuse sous tension avec le commutateur principal. Relâcher les deux touches ; il en sultera
l'exécution d'un test de l'affichage pendant environ 5 secondes. Tous les indicateurs appartront simultament
(voir la figure 29).
Figure. 29
ATTENTION :
L'utilisateur devra introduire le programme conformément à la description du paragraphe 4.1.1 afin de
modifier les glages des paragraphes 4.1.2 - 4.1.7. Après avoir mémorisé les glages, il est possible
de revenir au mode de programmation normale en mettant hors tension la centrifugeuse pendant un
court moment.
Tous les réglages modifiés doivent être confirmés avec la touche "Start"(9). Le mot "Store"(12) appart sur
l'affichage "RPM | RCF"(A-1) - Seules les présélections sont valides !
Figure. 30
4.1.2 Indication de la température
Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et
appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" appart.
Sélectionner maintenant la lettre "C" avec le bouton de réglage (1). Ceci entraîne l'apparition des mots
"CELSI/temp" sur l'affichage "RPM | RCF" (A-1). En appuyant sur la touche "RPM | RCF" (4), le mot "CELSI"
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-15
clignote et il devient possible de modifier l'affichage en degrés Fahrenheit "FAREN" avec le bouton de réglage (1)
(voir la figure 31).
Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en
mettant hors tension la centrifugeuse pendant un court moment.
Figure. 31
4.1.3 Signal Marche/Arrêt
Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et
appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" clignote.
Sélectionner maintenant la lettre "L" avec le bouton de glage (1). Ceci entraîne l'apparition des mots "On
Sound" sur l'affichage "RPM | RCF" (4). En appuyant maintenant sur la touche "RPM | RCF" (4), le mot "On"
clignote et il devient possible de couper le son avec le bouton de réglage (1) (voir la figure 32).
Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en
mettant hors tension la centrifugeuse pendant un court moment.
Figure. 32
4.1.4 Volume de présélection du signal sonore
Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et
appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" clignote.
Sélectionner maintenant la lettre "U" avec le bouton de réglage (1). Ceci entraîne l'apparition des mots "Vol=0-
9/Sound" sur l'affichage "RPM | RCF" (A-1). Après avoir appuyé sur la touche "RPM | RCF" (4), l'utilisateur peut
régler le volume désiré entre 0 (faible) et 9 (fort) avec le bouton de glage (1) (voir la figure 33).
Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en
mettant hors tension la centrifugeuse pendant un court moment.
Figure. 33
FR-16 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4.1.5 Sélection de la sonori du signal sonore de fin de centrifugation
Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et
appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" clignote.
Sélectionner maintenant la lettre "G" avec le bouton de glage (1). Ceci entraîne l'apparition des mots
"SonGo/Sound" sur l'affichage "RPM | RCF" (A-1). Après avoir appuyé sur la touche "RPM | RCF" (4), il est
possible de sélectionner une sonorité avec le bouton de glage (1). (voir la figure 34)
Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en
mettant hors tension la centrifugeuse pendant un court moment.
Figure. 34
4.1.6 Marche/Arrêt du son du clavier
Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et
appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" clignote.
Sélectionner maintenant la lettre "B" avec le bouton de glage (1). Ceci entraîne l'apparition des mots
"ON/BEEP " sur l'affichage "RPM | RCF" (A-1). Après avoir appuyé sur la touche "RPM | RCF" (4), l'utilisateur
peut mettre le son du clavier sur Marche ou sur Arrêt avec le bouton de glage (1). (voir la figure 38)
Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en
mettant hors tension la centrifugeuse pendant un court moment.
Figure. 35
4.1.7 Appel des données de fonctionnement
ATTENTION :
Ceci ne doit être alisé que par des utilisateurs avertis ou par des techniciens de maintenance.
Dans le mode « Réglages de base », il est possible de rappeler les dones de fonctionnement de la
centrifugeuse. Veuillez poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode
de programme. Appuyer sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service"
clignote.
Grâce au bouton de glage (1), il est possible d'accéder à différentes informations :
A = démarrages précédents de la centrifugeuse
H = nombre d'heures de fonctionnement
S = version du logiciel
r = logiciel de conversion
E = liste des messages d'erreur
h = durée de fonctionnement du moteur
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-17
La liste des 99 derniers messages d'erreur peut être consultée en appuyant sur la touche "RPM | RCF" (4) et en
faisant filer la liste avec le bouton de glage (1). Les codes d'erreurs apparaissent sur l'affichage "RPM | RCF"
(A-1). Se reporter au "Tableau 6 : messages d'erreur (voir ANNEXE).
Pour revenir au mode de programmation normale, mettre la centrifugeuse hors tension pendant un court moment.
Figure. 36
5. MAINTENANCE
5.1 Maintenance et nettoyage
5.1.1 Entretien néral :
L’entretien de la centrifugeuse est limité au nettoyage du rotor, de la cuve du rotor et de ses accessoires, et à la
lubrification régulière des boulons d'insertion de rotor pour un rotor horizontal (le cas échéant).
Le lubrifiant préconisé est «High TEF Oil»
Les lubrifiants contenant du disulfure de molybdène et du graphite ne sont pas autorisés.
Prêter particulièrement attention aux pièces en aluminium anodisé. La rupture de rotors peut être provoquée par de
ts légères tériorations.
Si des rotors, des godets ou des supports de tubes entrent en contact avec des substances corrosives, les
surfaces concernées doivent être nettoyées avec pcaution.
Voici par exemple quelques substances corrosives : alcali, solution savonneuse alcaline, amines alcalines, acides
concentrés, solution contenant des métaux lourds, solvants chlorés sans eau, solutions salines comme de l'eau
salée, du phénol et des hydrocarbures halogénés.
Nettoyage des unis, des rotors et des accessoires :
- Mettre l'appareil hors tension et le connecter de l'alimentation électrique avant de commencer un nettoyage ou
une désinfection. Ne pas verser de liquides à l'intérieur de l'enceinte.
- Ne pulvériser du désinfectant sur l'appareil.
- Un nettoyage complet est non seulement cessaire à titre d'hygiène, mais aussi pour éviter une corrosion due à
la pollution.
- Afin d'éviter d'endommager des pièces anodisées comme les rotors, les plaques de duction, etc., seuls des
tergents à pH neutre (pH 6-8) peuvent être utilisés pour le nettoyage. Des produits de nettoyage alcalins (pH >
8) ne doivent pas être utilisés.
- Après le nettoyage, s'assurer que toutes les pièces sont bien séchées soit manuellement, soit dans une armoire à
air chaud (température maximum + 50 °C).
- Il est nécessaire de revêtir gulièrement les pièces en aluminium anodisé avec de l'huile anti-corrosion afin
d'augmenter leur due de vie et de réduire les possibilités de corrosion.
- De la condensation peut se for mer en psence d'humidité ou si les échantillons ne sont pas hermétiquement
scellés. La condensation doit être gulièrement éliminée de la cuve du rotor avec un chiffon doux.
La procédure de maintenance doit être répétée toutes les 10 à 15 centrifugations, ou au moins une fois
par semaine.
- Connecter l'unité à son alimentation après avoir complètement séc l'équipement.
FR-18 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
- Aucune désinfection ne doit être effectuée avec des rayons ultraviolets, bêta et gamma, ou avec toute autre
énergie rayonnante élevée.
- Les rotors métalliques peuvent être passés à l'autoclave.
- Le couvercle du rotor et les adaptateurs peuvent aussi être passés à l'autoclave (maxi. U21 °C, 20 min).
- Les supports de tubes sont fabriqués en PP et ne peuvent pas être passés à l'autoclave à 134 °C.
5.1.2 Nettoyage et désinfection de l'unité
1. Ouvrir le couvercle avant de mettre l'unité hors tension. La déconnecter de son alimentation.
2. Dévisser l'écrou du rotor en tournant la clé du rotor dans le sens anti-horaire.
3. Démonter le rotor.
4. Nettoyer et désinfecter l'unité et la cuve du rotor en utilisant le nettoyant mention ci-dessus.
5. Nettoyer toutes les surfaces accessibles du dispositif et de ses accessoires, notamment le cordon électrique,
avec un chiffon humide.
6. Laver abondamment les joints en caoutchouc et la cuve du rotor avec de l'eau.
7. Frotter les joints en caoutchouc sec avec du talc ou de la glycérine afin d'éviter qu'ils ne deviennent cassants.
D'autres composants de l'appareil, comme le verrouillage du couvercle, l'arbre du moteur et le rotor, ne doivent
pas être graissés.
8. Sécher l'arbre du moteur avec un chiffon doux, sec et non pelucheux.
9. Vérifier l'unité et les accessoires à la recherche de détériorations.
S'assurer que la centrifugeuse est hors tension et déconnecter l'unité de l'alimentation électrique. Éliminer ensuite
la poussière adrant aux fentes d'ration dans la centrifugeuse avec une brosse souple. Ceci doit être alisé au
moins tous les six mois.
5.1.3 Nettoyage et désinfection du rotor
1. Nettoyer et désinfecter les rotors, les couvercles des rotors ainsi que les adaptateurs avec le nettoyant
mentionné ci-dessus.
2. Utiliser un goupillon pour nettoyer et sinfecter les alésages du rotor.
3. Rincer les rotors, le couvercle du rotor et l'adaptateur à l'eau claire. Particulièrement les passages perforés des
rotors angulaires.
4. Pour sécher les rotors et les accessoires, les installer sur une serviette. Positionner les rotors angulaires avec
les alésages vers le bas.
5. Sécher le cône du rotor avec un chiffon doux, sec et non pelucheux, tout en recherchant d'éventuelles
tériorations. Ne pas graisser le cône du rotor.
5.1.4 sinfection des rotors en aluminium
Si des produits infectieux se pandent dans la centrifugeuse, le rotor et la cuve du rotor doivent être désinfectés
immédiatement aps la centrifugation. Les rotors peuvent être passés à l'autoclave à une température maximum
de 121 °C.
5.1.5 sinfection des rotors PP
Autoclave
Durée recommandée de passage à l'autoclave : 15 20 min à 121 °C (1 bar)
ATTENTION :
La durée de stérilisation de 20 minutes ne doit pas être dépassée. Des stérilisations tées
provoqueront la diminution de la sistance canique des matériaux en plastique
Avant de passer le rotor PP et l'adaptateur à l'autoclave, ils doivent être bien nettoyés afin d'éviter de
« cuire » les sidus de saletés.
Ne pas tenir compte des effets de certains résidus chimiques sur les matériaux en plastique aux temratures
ambiantes. Mais avec les températures élevées en autoclave, ces résidus peuvent corroder et détruire le plastique.
Les objets doivent être abondamment rincés à l'eau distillée après le nettoyage mais avant de les passer à
l'autoclave. Les résidus de liquide de nettoyage peuvent provoquer des fissures, un blanchissement et des taches.
Stérilisation au gaz
Adaptateurs, bouteilles et rotors sont stérilisables au gaz avec de l'oxyde d'éthylène. S'assurer de laisser les
articles à l'air libre après la stérilisation et avant de les utiliser.
ATTENTION :
La température pouvant augmenter pendant la stérilisation, les rotors, les adaptateurs et les bouteilles
ne doivent pas être fermés et doivent être complètement dévissés.
Stérilisation chimique
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-19
Bouteilles, adaptateurs et rotors peuvent être traités avec des désinfectants liquides usuels.
ATTENTION :
Avant d'appliquer toute autre méthode de nettoyage ou de contamination que celle recommandée par
le fabricant, contacter ce dernier afin d'être assuré qu'elle ne tériorera pas
l'unité ou le rotor.
5.1.6 Bris de verre
Avec des valeurs d'accélération élevée, le taux de bris de tubes en verre augmente. Des éclats de verre doivent
être immédiatement éliminés du rotor, des godets, des adaptateurs et de la cuve du rotor elle-même. De fins éclats
de verre rayeront et endommageront le revêtement protecteur de surface d'un rotor. Si des éclats de verre restent
dans la cuve du rotor, une fine poussière métallique s'accumulera en raison de la circulation d'air. Cette ts fine
poussre métallique noire polluera fortement la cuve du rotor, le rotor, les godets et les échantillons.
Remplacer au besoin les adaptateurs, les tubes et les accessoires afin d'éviter d'amplifier les tériorations.
Vérifier régulièrement les alésages du rotor à la recherche de sidus et de détériorations.
ATTENTION :
Vérifier les spécifications particulières des tubes avec le fabricant.
Durée d'utilisation des rotors, des godets et des accessoires
Les rotors et le couvercle de rotor en aluminium ou en acier inoxydable ont une durée opérationnelle de 7 ans au
maximum après la première utilisation.
Les bouchons et les couvercles de rotor transparents fabriqs en PC ou en PP ainsi que les rotors, les supports
de tubes et les adaptateurs en PP ont une durée de fonctionnement maximum de 3 ans après la première
utilisation.
Conditions de la durée de fonctionnement : Utilisation adéquate, protections en place et entretien recomman.
6. DÉPANNAGE
6.1 Message d'erreur : Causes/solutions
Les messages d'erreur sont pertoriés afin d'aider à localiser plus rapidement de possibles erreurs.
Le diagnostic de référence dans ce chapitre peut ne pas toujours correspondre à la situation dans la mesure les
erreurs et les solutions proposées sont toriques.
Veuillez nous informer de toutes les sortes d'erreurs se produisant qui ne sont pas répertoriées dans ce chapitre.
C'est grâce à vos informations en retour que nous sommes en mesure d'améliorer ce manuel de fonctionnement.
Nous vous remercions dès maintenant de votre aide.
6.2 Étude des messages d'erreur possibles et leurs solutions
6.2.1 Liration du couvercle pendant une panne d'alimentation (libération en cas d'urgence)
En cas de panne d'alimentation ou de faut de fonctionnement, le couvercle de la centrifugeuse peut être ouvert
manuellement afin récupérer les échantillons.
Pour les modèles FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (moteur de verrouillage entraîné)
Poursuivre de la manière suivante :
1. Mettre la centrifugeuse hors tension, brancher le cordon d'alimentation et attendre que le rotor soit à l'arrêt
complet (ce qui peut prendre plusieurs minutes)
2. Sur le côté gauche de l'enceinte de la centrifugeuse se trouve une butée en plastique. Retirer la butée derrière
laquelle se trouve un écrou hexagonal.
3. Utiliser la cpolygonale incluse, l'introduire dans l'orifice et serrer la cpolygonale avec l'écrou hexagonal
(voir la figure 37).
4. Faire ensuite tourner la clé polygonale vers la droite (dans le sens horaire) jusqu'à la limite.
ATTENTION :
a) Ne tourner que jusqu'à la limite sans serrer lcrou.
b) Il est dorénavant possible d'ouvrir le couvercle de
la centrifugeuse.
c) Remettre la centrifugeuse sous tension pour reprendre le travail.
FR-20 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figure. 37
Pour 5714
Procéder comme suit (cf. Figure 38):
ATTENTION:
Arrêter la centrifugeuse et brancher le cordon électrique, attendre que le rotor soit à l’arrêt
(cela peut prendre plusieurs minutes).
Sur le côté droit, il y a un bouchon en plastique (Figure 38). Ôter le bouchon, raccor au verrou
du capot, horizontalement du corps, jusqu’à ce que le capot de la centrifugeuse soit ouvert.
Ouvrir maintenant le capot de la centrifugeuse.
Figure 38
6.2.2 Description du sysme de messages d'erreur
Le message d'erreur "error" (M11) appart sur l'affichage "Time/Prog" (A-3) (voir la figure 39). Des informations
taillées sur les messages d'erreur possibles se trouvent sur le « tableau 6 : messages d'erreur" (voir Annexe).
Figure. 39
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-21
7. RÉCEPTION DES CENTRIFUGEUSES POUR RÉPARATION
Danger pour la santé par des équipements, des rotors et des accessoires contamis.
En cas de retour d'une centrifugeuse pour paration auprès du fabricant, prendre note de ce qui suit :
La centrifugeuse doit être décontaminée et nettoyée avant expédition afin de protéger les personnes,
l'environnement et le matériel.
Certificat de décontamination pour le retour de matériel (voir ANNEXE)
Le fabricant se serve le droit de ne pas accepter de centrifugeuses contaminées.
De plus, toutes les dépenses de nettoyage et de désinfection des unités seront bitées du compte du client
8. TRANSPORT, STOCKAGE ET MISE AU REBUT
8.1 Transport
Avant le transport, extraire le rotor.
Ne transporter cette unité que dans son emballage d'origine.
Utiliser une aide au transport pour une exdition sur de longues distances
afin de bloquer l'arbre du moteur.
Temrature de
l'air
Humidité
relative
Pression d'air
néralités sur le
transport
-25 à 60 °
10 à 75 %
30 à 106 kPa
8.2 Stockage
Pendant le stockage de la centrifugeuse, les conditions environnementales
suivantes doivent être observées :
Temrature de l'air
Humidité relative
Pression d'air
dans l'emballage de
transport
-25 à 55 °
10 à 75 %
70 à 106 kPa
8.3 Transport, installation, transfert et mise au rebut de la centrifugeuse
FC5714/FC5718/FC5718R/FC5816/FC5816R
Ces instructions viennent en complément des instructions générales précédentes du chapitre 8 et ne les
remplacent pas.
8.3.1 Transport
Transporter l'équipement dans son emballage d'origine.
La centrifugeuse doit toujours être transportée par deux personnes.
Utiliser une aide au transport pour une exdition sur de plus longues
distances.
8.3.2 Installation
La centrifugeuse doit toujours être transportée par deux personnes.
Ouverture du carton et extraction de l'équipement.
1. Couper la bande adsive.
2. Ouvrir les 4 rabats du carton.
3. Enlever les accessoires.
4. Placer les mains sous l'équipement et soulever la centrifugeuse hors du
carton avec une autre personne.
FR-22 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Placer l'équipement sur une table de laboratoire stable, horizontale et anti-vibratoire
1. Enlever le matériau de protection à l'avant et à l'arrière.
2. Enlever le manchon en plastique.
3. Respecter une distance minimum de 30 cm sur les côtés (dispositifs adjacents)
et à l'arrière (mur).
4. Installer l'équipement dans un endroit bien ré, protégé de la lumière directe
du soleil pour éviter toute surchauffe.
Connexion de l'équipement
1. Après l'installation, attendre quatre heures avant de brancher la centrifugeuse afin d'éviter d'endommager le
compresseur.
2. Vérifier que la tension et la fréquence du réseau électrique correspondent aux
scifications sur la plaque de l'équipement (voir l'arrière du dispositif), puis
connecter l'équipement à l'alimentation électrique.
Enlever le matériau de protection de la cuve du rotor
1. Brancher l'équipement sur l'interrupteur d'alimentation.
2. Ouvrir le couvercle de la centrifugeuse à l'aide du bouton d'ouverture.
4. Enlever le matériau de protection utilisé pour le transport.
5. Placer le rotor à la verticale sur l'arbre du moteur.
6. Tourner l'écrou du rotor dans le sens horaire à l'aide de la clé de rotor jusqu'à
ce que l'écrou soit bien serré.
L'équipement est maintenant prêt à l'utilisation
Conserver l'emballage et tous les matériaux de protection en cas d'expédition
ultérieure de l'équipement.
8.3.3 Emballage
Emballer la centrifugeuse dans l'ordre inverse.
8.3.4 Transfert de l'équipement
Lors du passage de l'équipement à des tiers, s'il vous plaît assurez-vous
d'inclure également ce mode d'emploi.
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-23
9. DONNEES TECHNIQUES
9.1 Spécifications
9.1.1 Centrifugeuse FC5714
Modèle
FC5714
Plage des vitesses
200 rpm -14000 rpm;10 rpm/set
Maximum RCF
18624 x g;10 x g/set
Capacimaximum (rotor)
4x100ml
Plage de température
Air froid
Due de fonctionnement
10 sec à 99 h 59 min 59 sec ou en continu
Niv eau du bruit (en fonction du
rotor)
63 ± 2 dB(A)
Densiacceptable à la v itesse
maximum
1.2 g/ml
Énergie cinétique acceptable
5595 Nm
Connexion électrique CA
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Variations de tension
± 10 %
± 10 %
Consommation électrique
1.3 A
2.4 A
Consommation
240 W
300 W
Dimensions (L × P × H)
362 x 493 x 330 mm
14.3 x 19.4 x 13.0 in
Poids net (sans rotor)
30 kg
66 lb
Dimensions d'exdition (L × P ×
H)
580 x 490 x 460 mm
22.8 x 19.3 x 18.1 in
Poids à l'exdition (sans rotor)
32.5 kg
72 lb
Conditions ambiantes (EN/IEC 61010-1)
Environnement
pour utilisation à l’intérieur uniquement
Altitude
Utilisation jusqu'à une altitude maximum de 2000 mètres
Température ambiante
2°C jusqu'à 35 °C
Humidité relativ e maximum
Humidité relative maximum de 80 % pour des
températures jusqu 31 °C,
avec diminution linéaire jusqu'à 50 % de l'humidité
relative à 35 °C.
Catégorie de surtension
(IEC 60364-4-443)
II
Niv eau de contamination
2
Classe de protection
I
Inadaptée à une utilisation dans des endroits dangereux.
FR-24 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.2 Centrifugeuse FC5718
Modèle
FC5718
Plage des vitesses
200 rpm -18000 rpm;10 rpm/set
Maximum RCF
23542 x g;10 x g/set
Capacimaximum (rotor)
4x100ml
Plage de température
Air froid
Due de fonctionnement
10 sec à 99 h 59 min 59 sec ou en continu
Niv eau du bruit (en fonction du
rotor)
60 ± 2 dB(A)
Densiacceptable à la v itesse
maximum
1.2 g/ml
Énergie cinétique acceptable
16672 Nm
Connexion électrique CA
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Variations de tension
± 10 %
± 10 %
Consommation électrique
2.0 A
4.0 A
Consommation
455 W
475 W
Dimensions (L × P × H)
408 x 499 x 351 mm
16.1 x 19.7 x 13.8 in
Poids net (sans rotor)
43 kg
95 lb
Dimensions d'exdition (L × P ×
H)
650 x 520 x 490 mm
25.6 x 20.5 x 19.3 in
Poids à l'exdition (sans rotor)
53 kg
117 lb
Conditions ambiantes (EN/IEC 61010-1)
Environnement
pour utilisation à l’intérieur uniquement
Altitude
Utilisation jusqu'à une altitude maximum de 2000 mètres
Température ambiante
2 °C jusqu'à 35 °C
Humidité relativ e maximum
Humidité relative maximum de 80 % pour des
températures jusqu 31 °C,
avec diminution linéaire jusqu'à 50 % de l'humidité
relative à 35 °C.
Catégorie surtension
(IEC 60364-4-443)
II
Niv eau de contamination
2
Classe de protection
I
Inadaptée à une utilisation dans des endroits dangereux.
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-25
9.1.3 Centrifugeuse FC5718R
Modèle
FC5718R
Plage des vitesses
200 rpm -18000 rpm;10 rpm/set
Maximum RCF
23542 x g;10 x g/set
Capacimaximum (rotor)
4x100ml
Plage de température
-20~40
Due de fonctionnement
10 sec à 99 h 59 min 59 sec ou en continu
Niv eau du bruit (en fonction du
rotor)
60 ± 2 dB(A)
Densiacceptable à la v itesse
maximum
1.2 g/ml
Énergie cinétique acceptable
25111 Nm
Connexion électrique CA
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Variations de tension
± 10 %
± 10 %
Consommation électrique
3.0 A
6.0 A
Consommation
660 W
660 W
Dimensions (L × P × H)
407 x 731 x 359 mm
16.0 x 28.8 x 14.1 in
Poids net (sans rotor)
60 kg
132 lb
Dimensions d'exdition (L × P ×
H)
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
Poids à l'exdition (sans rotor)
77 kg
170 lb
Conditions ambiantes (EN/IEC 61010-1)
Environnement
pour utilisation à l’intérieur uniquement
Altitude
Utilisation jusqu'à une altitude maximum de 2000 mètres
Température ambiante
2 °C jusqu'à 35 °C
Humidité relativ e maximum
Humidité relative maximum de 80 % pour des
températures jusqu 31 °C,
avec diminution linéaire jusqu'à 50 % de l'humidité
relative à 35 °C.
Catégorie surtension
(IEC 60364-4-443)
II
Niv eau de contamination
2
Classe de protection
I
Inadaptée à une utilisation dans des endroits dangereux.
FR-26 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.4 Centrifugeuse FC5816
Modèle
FC5816
Plage des vitesses
200 rpm -15000 rpm;10 rpm/set
Maximum RCF
21379 x g;10 x g/set
Capacimaximum (rotor)
6 x 250ml
Plage de température
Air froid
Due de fonctionnement
10 sec à 99 h 59 min 59 sec ou en continu
Niv eau du bruit (en fonction du
rotor)
61 ± 2 dB(A)
Densiacceptable à la v itesse
maximum
1.2 g/ml
Énergie cinétique acceptable
34363 Nm
Connexion électrique CA
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Variations de tension
± 10 %
± 10 %
Consommation électrique
2.4 A
4.2 A
Consommation
530 W
520 W
Dimensions (L × P × H)
446 x 538 x 354 mm
17.6 x 21.2 x 13.9 in
Poids net (sans rotor)
52 kg
115 lb
Dimensions d'exdition (L × P ×
H)
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
Poids à l'exdition (sans rotor)
77 kg
170 lb
Conditions ambiantes (EN/IEC 61010-1)
Environnement
pour utilisation à l’intérieur uniquement
Altitude
Utilisation jusqu'à une altitude maximum de 2000 mètres
Température ambiante
2 °C jusqu'à 35 °C
Humidité relativ e maximum
Humidité relative maximum de 80 % pour des
températures jusqu 31 °C,
avec diminution linéaire jusqu'à 50 % de l'humidité
relative à 35 °C.
Catégorie surtension
(IEC 60364-4-443)
II
Niv eau de contamination
2
Classe de protection
I
Inadaptée à une utilisation dans des endroits dangereux.
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-27
9.1.5 Centrifugeuse FC5816R
Modèle
FC5816R
Plage des vitesses
200 rpm -16000 rpm;10 rpm/set
Maximum RCF
24325 x g;10 x g/set
Capacimaximum (rotor)
6 x 250ml
Plage de température
-20~40
Due de fonctionnement
10 sec à 99 h 59 min 59 sec ou en continu
Niv eau du bruit (en fonction du
rotor)
63 ± 2 dB(A)
Densiacceptable à la v itesse
maximum
1.2 g/ml
Énergie cinétique acceptable
34363 Nm
Connexion électrique CA
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Variations de tension
± 10 %
± 10 %
Consommation électrique
3.7 A
7.8 A
Consommation
785 W
850 W
Dimensions (L × P × H)
723 x 538 x 354 mm
28.5 x 21.2 x 13.9 in
Poids net (sans rotor)
77 kg
170 lb
Dimensions d'exdition (L × P ×
H)
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
Poids à l'exdition (sans rotor)
87 kg
192 lb
Conditions ambiantes (EN/IEC 61010-1)
Environnement
pour utilisation à l’intérieur uniquement
Altitude
Utilisation jusqu'à une altitude maximum de 2000 mètres
Température ambiante
2 °C jusqu'à 35 °C
Humidité relativ e maximum
Humidité relative maximum de 80 % pour des
températures jusqu 31 °C,
avec diminution linéaire jusqu'à 50 % de l'humidité
relative à 35 °C.
Catégorie surtension
(IEC 60364-4-443)
II
Niv eau de contamination
2
Classe de protection
I
Inadaptée à une utilisation dans des endroits dangereux.
FR-28 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.6 Centrifuge FC5916
Modèle
FC5916
Plage de vitesse
200 tr/min -16000 tr/min;10 tr/min/set
ACR max.
24325 x g;10 x g/set
Capacimaxi(Rotor)
4x750ml
Plage de température(N/A)
Refroidissement par air
Due de fonctionnement
10 sec à 99 h 59 min 59 sec ou en continu
Niv eau sonore (selon le rotor)
63 ± 2 dB(A)
Densiadmissible à la v itesse
maximale
1.2 g/ml
Énergie cinétique admissible
60629 Nm
Raccordement au seau AC
230 V ~ 50/60 Hz
120 V ~
50/60 Hz
Fluctuation de la tension
± 10 %
± 10 %
Consommation de courant
2.8 A
5.6 A
Consommation d'énergie
640 W
680 W
Dimensions (L × P × H)
540 x 670 x 390 mm
21.3 x 26.4 x 15.4 in
Poids net (sans rotor)
85 kg
187 lb
Dimensions d'exdition (L × P ×
H)
780 x 670 x 590 mm
30.7 x 26.4 x 23.3 in
Poids à l'exdition (sans rotor)
98 kg
216 lb
Conditions ambiantes (EN/IEC 61010-1)
Environnement
Pour usage intérieur seulement
Altitude
Utilisation jusqu'à une altitude de 2000 m
Température ambiante
2°C jusqu 35 °C
Humidité relativ e max.
Humidité relative max. 80 % pour des températures
jusqu'à 31°C,
diminuant linéairement jusqu 50 % d'humidité relative
jusqu'à 35°C.
Catégorie de surtension (IEC
60364-4-443)
II
Degré de contamination
2
Classe de protection
I
Ne convient pas à une utilisation
dans des environnements
dangereux.
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-29
9.1.7 Centrifuge FC5916R
Modèle
FC5916R
Plage de v itesse
200 rpm -16000 rpm;10 rpm/set
ACR max.
26331 x g;10 x g/set
Capacimaxi(Rotor)
4x750ml
Plage de température(Digitale)
-20~40
Due de fonctionnement
10 sec à 99 h 59 min 59 sec ou en continu
Niv eau sonore (selon le rotor)
63 ± 2 dB(A)
Densiadmissible à la v itesse
maximale
1.2 g/ml
Énergie cinétique admissible
54458 Nm
Raccordement au seau AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Fluctuation de la tension
± 10 %
± 10 %
Consommation de courant
7.2 A
20 A
Consommation d'énergie
1630 W
1750 W
Dimensions (L × P × H)
730 x 670 x 390 mm
28.7 x 26.4 x 15.4 in
Poids net (sans rotor)
118 kg
260 lb
Dimensions d'exdition (L × P ×
H)
900 x 750 x 560 mm
40.0 x 29.5 x 22.0 in
Poids à l'exdition (sans rotor)
137 kg
137 kg
Conditions ambiantes (EN/IEC 61010-1)
Environnement
Pour usage intérieur seulement
Altitude
Utilisation jusqu'à une altitude de 2000 m
Température ambiante
2°C jusqu 35 °C
Humidité relativ e max.
Humidité relative max. 80 % pour des températures
jusqu'à 31°C,
diminuant linéairement jusqu 50 % d'humidité relative
jusqu'à 35°C.
Catégorie de surtension (IEC
60364-4-443)
II
Degré de contamination
2
Classe de protection
I
Ne convient pas à une utilisation dans des env ironnements dangereux.
FR-30 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.2 Schémas et dimensions
D
H
W
W
H
D
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-31
Modèle
L (mm / in.)
P (mm / in.)
H (mm / in.)
FC5714
362 / 14.3
493 / 19.4
330 / 13.0
FC5718
408 / 16.1
499 / 19.7
351 / 13.8
FC5718R
408 / 16.1
731 / 28.8
359 / 14.1
FC5816
446 / 17.6
538 / 21.2
354 / 13.9
FC5816R
723 / 28.5
538 / 21.2
354 / 13.9
FC5916
540 / 21.3
670 / 26.4
390 / 15.4
FC5916R
730 / 28.7
670 / 26.4
390 / 15.4
10. CONFORMITÉ
La conformi aux normes suivantes est indiquée par le marquage correspondant sur le produit.
Marquage
Standard
Ce produit est conforme aux normes harmonisées applicables des directives
européennes 2011/65/UE (RoHS), 2014/30/UE (CEM) et 2014/35/UE (LVD).
La déclaration de conformité de l'UE est disponible en
ligne à www.ohaus.com/ce.
Mise au rebut
Conformément à la directive euroenne 2012/19/UE sur les déchets
d’équipements électriques et électroniques (WEEE), cet appareil ne doit pas être
mis au rebut avec les ordures ménagères. Ceci est également valable pour les
pays en dehors de l’UE, selon leurs scifications particulières.
Veuillez mettre au rebut ce produit conformément à la réglementation locale au
point de collecte spécifpour les équipements électriques et électroniques.
En cas de questions, veuillez contacter l’autorité responsable ou le distributeur
auprès duquel cet appareil a été acheté.
Si cet appareil change de propriétaire (pour des raisons personnelles ou
professionnelles), cette consigne doit lui être communiquée.
Pour les instructions de rejet et recyclage en Europe, merci de consulter le site
www.ohaus.com/weee.
Nous vous remercions de votre contribution à la protection de l’environnement.
FR-32 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11. ANNEXE
TABLEAU 1 : POIDS NET AUTORISE
TABLEAU 2 : TEMPERATURES LES PLUS FAIBLES A LA VITESSE MAXIMUM
TABLEAU 3 : VITESSE ET VALEURS RCF MAXI POUR LES ROTORS AUTORISES
TABLEAU 4 : DUREES D'ACCELERATION ET DE DECELERATION
TABLEAU 5 : MESSAGES D'ERREUR
TABLEAU 6: CORRECTION DU RAYON
TABLEAU 7 : FORMULAIRE DE RACHAT/CERTIFICAT DE DECONTAMINATION
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-33
11.1 Table 1: POIDS NET AUTORISE
Numé ro de rotor
Description du rotor
Poids net autori
Vitesse maxi ( x1000 tr/min)
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
16.0
16.0
30314820
Rotor, Sw ing out, 4x250ml, ID
4x 536 g
4.5
4.5
30314821
Rotor, Angle, 6x 250ml FB, ID
6x 355 g
8.0
8.0
8.0
8.0
30314822
Rotor, Sw ing out, 4x100ml, ID
4x 340 g
4.5
4.5
4.5
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
4x 465 g
4.0
5.0
5.0
30314824
Rotor, Sw ing out, 2x3MTP, ID
2x 310 g
4.5
4.5
4.5
4.5
4.5
4.5
4.5
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
6x 140 g
11.0
13.5
30314826
Rotor, Angle, 6x85ml RB, ID
6x 140 g
9.0
9.0
11.0
13.0
11.0
11.0
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
4x 140 g
12.0
12.0
12.0
12.0
15.0
15.0
30314828
Rotor, Sw ing out, 16x 50ml, ID
4x 557 g
4.5
4.5
FR-34 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314829
Rotor, Angle, 10x50ml FA, ID
10x 74 g
7.5
7.5
9.0
10.5
10.0
10.0
30314830
Rotor, Angle, 6x50ml RB/FA, ID
6x 72 g
6.0
6.0
6.0
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
6x 94 g
12.0
12.0
13.0
13.0
13.0
13.0
30314832
Rotor, Angle, 30x15ml RB/FA, ID
30x 32 g
4.5
4.5
4.5
4.5
4.5
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
20x 18 g
12.0
12.0
12.0
12.0
12.0
12.0
30314834
Rotor, Angle, 12x15ml RB/FA, ID
12x 25 g
6.0
6.0
6.0
30314835
Rotor, Angle, 44x 1.5/2.0ml, ID
44x 3.8 g
13.5
13.5
15.0
16.0
16.0
16.0
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
30x 3.4 g
12.0
13.0
14.0
15.0
15.0
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
15.0
15.0
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
12x 3.4 g
18.0
18.0
30314840
Rotor, Angle, 64x0.5ml, ID
64x 2.3 g
13.5
13.5
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
4x 3.5 g
15.0
15.0
15.0
15.0
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
4x 995 g
4.0
4.0
30553086
Rotor Angle 4x 500ml D69mm ID
4x 708 g
8.0
8.0
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-35
11.2 Table 2: TEMPERATURES LES PLUS FAIBLES A LA VITESSE MAXIMUM
Numé ro de rotor
Description du rotor
Modè le
Vitesse maximale
N-max
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314821
Rotor, Angle, 6x250ml FB, ID
FC5916R
8000 rpm
-1 ° C
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5718R
4500 rpm
-7 ° C
FC5816R
4500 rpm
-7 ° C
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5718R
5000 rpm
-3 ° C
FC5816R
5000 rpm
-3 ° C
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5718R
4500 rpm
-6 ° C
FC5816R
4500 rpm
-6 ° C
FC5916R
4500 rpm
-8 ° C
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718R
13500 rpm
17 ° C
FC5816R
13500 rpm
17 ° C
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718R
9000 rpm
-5 ° C
FC5816R
9000 rpm
-5 ° C
FC5916R
13000 rpm
8 ° C
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718R
12000 rpm
5 ° C
FC5816R
12000 rpm
5 ° C
FC5916R
16000 rpm
7 ° C
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718R
7500 rpm
9 ° C
FC5816R
7500 rpm
9 ° C
FC5916R
10500 rpm
2 ° C
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5718R
6000 rpm
-8 ° C
FC5816R
6000 rpm
-8 ° C
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718R
12000 rpm
-5 °C
FC5816R
12000 rpm
-5 ° C
FC5916R
13000 rpm
-4 ° C
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5718R
4500 rpm
-9 ° C
FC5816R
4500 rpm
-9 ° C
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718R
12000 rpm
2 ° C
FC5816R
12000 rpm
2 ° C
FC5916R
12000 rpm
-2 ° C
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5718R
6000 rpm
-11.5 ° C
FC5816R
6000 rpm
-11.5 ° C
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718R
13500 rpm
1 ° C
FC5816R
13500 rpm
1 ° C
FC5916R
16000 rpm
3 ° C
30314836
Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable
FC5718R
14000 rpm
9 ° C
FC5816R
14000 rpm
9 ° C
FC5916R
15000 rpm
3 ° C
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718R
18000 rpm
-2 ° C
FC5816R
18000 rpm
-2 ° C
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718R
13500 rpm
2 ° C
FC5816R
13500 rpm
2 ° C
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718R
15000 rpm
-1 ° C
FC5816R
15000 rpm
-1 ° C
FC5916R
15000 rpm
-2 ° C
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916R
4500 rpm
4 ° C
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916R
8000 rpm
0 ° C
Toutes les indications de température se fèrent à une température ambiante de 23 ° C. En passant cette
valeur ou le rayonnement solaire direct vers la centrifugeuse, ces valeurs ne peuvent pas être maintenues.
FR-36 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.3 Table 3: VITESSE ET VALEURS RCF MAXI POUR LES ROTORS AUTORISES
Numé ro de rotor
Description du rotor
Modè le
Vitesse maximale
RPM
RCF
30304361
Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
4500
3780 x g
FC5816R
4500
3780 x g
30314821
Rotor, Angle, 6x250ml FB, ID
FC5816
8000
10016 x g
FC5816R
8000
10016 x g
FC5916
8000
10016 x g
FC5916R
8000
10016 x g
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5714
4500
3350 x g
FC5718
4500
3350 x g
FC5718R
4500
3350 x g
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
4000
2486 x g
FC5718
5000
3885 x g
FC5718R
5000
3885 x g
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5714
4500
2716 x g
FC5718
4500
2716 x g
FC5718R
4500
2716 x g
FC5816
4500
2716 x g
FC5816R
4500
2716 x g
FC5916
4500
2716 x g
FC5916R
4500
2716 x g
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
11000
13932 x g
FC5718R
13500
20984 x g
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
9000
10413 x g
FC5718R
9000
10413 xg
FC5816
11000
15555 x g
FC5816R
13000
21726 x g
FC5916
11000
15555 x g
FC5916R
13000
21726 x g
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718
12000
14809 x g
FC5718R
12000
14809 x g
FC5816
12000
14809 x g
FC5816R
12000
14809 x g
FC5916
15000
23140 x g
FC5916R
16000
26328 x g
30314828
Rotor, Sw ing out, 16x50ml, ID
FC5816
4500
3735 x g
FC5816R
4500
3735 x g
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-37
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718
7500
8174 x g
FC5718R
7500
8174 x g
FC5816
9000
11771 x g
FC5816R
10500
16022 xg
FC5916
10000
14532 x g
FC5916R
10500
16022 x g
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
12000
13522 x g
FC5718R
12000
13522 x g
FC5816
13000
15869 x g
FC5816R
13000
15869 x g
FC5916
13000
15869 x g
FC5916R
13000
15869 x g
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5714
4500
2830 x g
FC5718
4500
2830 x g
FC5718R
4500
2830 x g
FC5816
4500
2830 x g
FC5816R
4500
2830 x g
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
12000
15775 x g
FC5718R
12000
15775 x g
FC5816
12000
15775 x g
FC5816R
12000
15775 x g
FC5916
12000
15775 x g
FC5916R
12000
15775 x g
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718
13500
17113 x g
FC5718R
13500
17113 x g
FC5816
15000
21128 x g
FC5816R
16000
24039 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314836
Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable
FC5714
12000
15131 x g
FC5718
13000
17758 x g
FC5718R
14000
20595 x g
FC5916
15000
23643 x g
FC5916R
15000
23643 x g
FR-38 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718
18000
23542 x g
FC5718R
18000
23542 x g
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
13500
16298 x g
FC5718R
13500
16298 x g
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
15000
15343 x g
FC5718R
15000
15343 x g
FC5916
15000
15343 x g
FC5916R
15000
15343 x g
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916
4000
3434 x g
FC5916R
4500
4346 x g
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916
8000
10367 x g
FC5916R
8000
10367 x g
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-39
11.4 Table 4: DUREES D'ACCELERATION ET DE DECELERATION
N° d'article
Description du rotor
Modè le
Accé lé ration* Temps en
sec
Dé cé lé ration* Temps en
sec
niveau 0
niveau 9
niveau 0
niveau 9
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5714
200
25
230
35
FC5718
240
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
20
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
309
34
458
36
FC5816R
309
34
458
36
30314821
Rotor, Angle, 6x250ml FB, ID
FC5816
664
130
2906
92
FC5816R
664
130
2906
92
FC5916
573
66
1903
84
FC5916R
573
66
1903
84
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5714
110
14
170
19
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
110
14
170
17
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314824
Rotor, Sw ing out, 2x 3MTP, ID
FC5714
230
25
340
26
FC5718
160
20
360
15
FC5718R
160
20
360
15
FC5816
452
43
616
38
FC5816R
452
43
616
38
FC5916
249
27
488
23
FC5916R
249
27
488
23
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
400
40
960
40
FC5718R
500
60
1260
50
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
360
40
1050
40
FC5718R
360
40
1050
40
FC5816
697
85
2313
70
FC5816R
2313
70
1630
76
FC5916
463
48
1654
46
FC5916R
549
69
1307
67
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718
300
60
820
40
FC5718R
300
60
820
40
FC5816
506
60
1745
49
FC5816R
506
60
1745
49
FC5916
448
50
1251
45
FC5916R
448
50
1251
45
30314828
Rotor, Sw ing out, 16x50ml, ID
FC5816
311
34
387
36
FC5816R
311
34
387
36
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718
380
63
796
37
FC5718R
380
62
1069
36
FC5816
753
115
2395
72
FC5816R
753
115
2395
72
FC5916
480
60
1747
68
FC5916R
480
60
1747
68
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5714
101
11
206
14
FC5718
89
13
239
11
FC5718R
90
12
463
11
FR-40 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
446
48
1323
49
FC5816R
446
48
1323
49
FC5916
264
28
921
32
FC5916R
264
28
921
32
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5714
157
19
370
20
FC5718
160
15
380
10
FC5718R
160
15
380
10
FC5816
149
25
985
20
FC5816R
149
25
985
20
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
512
54
1439
47
FC5816R
512
54
1439
47
FC5916
305
32
988
37
FC5916R
305
32
988
37
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
102
11
167
14
FC5718
88
13
222
11
FC5718R
88
12
433
11
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718
196
22
204
18
FC5718R
196
22
204
18
FC5816
257
29
436
28
FC5816R
274
31
485
29
FC5916
233
25
283
30
FC5916R
233
25
500
25
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5714
250
27
280
34
FC5718
200
25
360
25
FC5718R
210
25
360
30
FC5916
221
23
561
30
FC5916R
221
23
561
30
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5714
200
22
230
35
FC5718
210
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
30
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718
240
25
210
20
FC5718R
240
25
210
20
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
150
17
170
12
FC5718R
150
17
170
12
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
130
16
130
12
FC5718R
130
16
130
12
FC5916
100
12
201
12
FC5916R
100
12
201
12
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916
483
47
1287
49
FC5916R
551
55
1501
54
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916
575
73
2317
82
FC5916R
575
73
2317
82
*Note: accelerates from 0 to Vmax; decelerates from Vmax to 0.
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-41
11.5 Tableau 5 : MESSAGES D'ERREUR
d'erreur :
Description
1
Survenue d'un dé sé quilibrage
2
Le capteur de dé sé quilibrage est dé fectueux
4
Le contacteur de dé sé quilibrage a é té activé pendant plus de
5 secondes
8
Le transpondeur dans le rotor est dé fectueux
11
Le capteur de tempé rature est dé fectueux
12
Surchauffe de la cuve
14
La variation de vitesse est trop importante entre deux mesures
FERMETURE
du couvercle
33
Ouverture du couvercle lorsque le moteur est en fonctionnement
34
Contact du couvercle dé fectueux
38
Le moteur du couvercle est bloqué
40
Les communications avec le convertisseur de fré quence sont
perturbé es pendant le dé marrage
41
Les communications avec le convertisseur de fré quence sont
perturbé es pendant l'arrê t
42
Court-circuit dans le convertisseur de fré quence
43
Tension trop faible sur le convertisseur de fré quence
44
Surtension sur le convertisseur de fré quence
45
Surchauffe du convertisseur de fré quence
46
Surchauffe du moteur
47
Surintensité du convertisseur de fré quence
48
Temporisation entre l'unité de cont le et le convertisseur de f quence
49
Autre erreur du convertisseur de fré quence
55
Survitesse
70
Temporisation entre le cont leur et l'interface RS232
99
Rotor non autorisé dans cette centrifugeuse
FAUX
Le rotor insé ré n'existe pas dans le programme
rotor
Le rotor n'est pas dé tecté
FR-42 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.6 Table 6: CORRECTION DU RAYON
Rotor
Description du
rotor
Godet/Ada
ptateur
Description du godet/adaptateur
Radius
(cm)
Correction
(cm)
Compatible
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle,
24x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314820
Rotor, Sw ing out,
4x 250ml, ID
Blank
NO Bucket/Adapter
16.7
30314903
Rack, 1x250ml D62mm FB, 2/pk
15.7
-1.0
30314904
Rack, 1x100ml D41mm RB, 2/pk
15.9
-0.8
30314905
Rack, 3x50ml D29mm RB, 2/pk
16.3
-0.4
30314906
Rack, 3x50ml D29mm FA, 2/pk
16.7
0.0
30314907
Rack, 1x50ml D34mm RB, 2/pk
15.8
-0.9
30314908
Rack, 2x50ml D29mm Rim, 2/pk
15.8
-0.9
30314909
Rack, 1x 50ml D29mm Rim, 2/pk
16.1
-0.6
30314910
Rack, 5x25ml D24.5mm RB, 2/pk
16.1
-0.6
30314911
Rack, 3x25ml D24.5mm Rim, 2/pk
15.5
-1.2
30314912
Rack, 9x15ml D17mm RB, 2/pk
16.3
-0.4
30314913
Rack, 7x15ml D17mm FA, 2/pk
16.4
-0.3
30314914
Rack, 7x15ml D17mm Rim, 2/pk
16.1
-0.6
30314915
Rack, 8x10ml D16mm RB, 2/pk
16.3
-0.4
30314916
Rack, 14x5-7ml D13mm RB, 2/pk
15.9
-0.8
30314917
Rack, 10x1.6-7ml D13mm Vac, 2/pk
15.9
-0.8
30304367
Rack, 8x4-10ml D16mm Vac, 2/pk
16.3
-0.4
30304368
Adapter, 9x 1.5/2.0ml D11mm, 2/pk
15.7
-1.0
30314821
Rotor, Angle,
6x 250ml FB, ID
Blank
NO Bucket/Adapter
14.1
30304369
Adapter, 8x 1.5ml D11mm, 2/pk
13.2
-0.9
30304370
Adapter, 5x 10ml D16mm RB, 2/pk
13.3
-0.8
30304371
Adapter, 4x 15ml D17mm FA, 2/pk
13.0
-1.1
30304372
Adapter, 2x 30ml D26mm RB, 2/pk
12.5
-1.6
30304373
Adapter, 1x 50ml D29mm RB, 2/pk
12.0
-2.1
30304374
Adapter, 1x 50ml D30mm FA, 2/pk
11.7
-2.4
30559412
Adapter 15x 1.5/2.0ml D11mm
MTBx 2 V1
30559413
Adapter 1x 250ml D60mm CBCx3 V1
11.8
-0.5
30559414
Adapter 1x 175/225ml D61mm FAx 2
V1
12.8
-2.3
30314822
Rotor, Sw ing out,
4x 100ml, ID
Blank
NO Bucket/Adapter
14.8
30314842
Rack, 1x100ml D46mm RB, 2/pk
14.1
-0.7
30314843
Rack, 1x85ml D38mm RB, 2/pk
14.1
-0.7
30314844
Rack, 1x50ml D34mm RB, 2/pk
14.2
-0.6
30314845
Rack, 1x50ml D29mm RB, 2/pk
14.2
-0.6
30314846
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.7
30314847
Rack, 1x50ml D29mm Rim, 2/pk
14.3
-0.5
30314848
Rack, 1x 30ml D25mm Rim, 2/pk
13.8
-1.0
30314849
Rack, 1x25ml D25mm RB, 2/pk
14.4
-0.4
30314850
Rack, 2x15ml D17mm FA, 2/pk
14.8
0.0
9.5
13.6
-4.6
-0.5
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-43
30314851
Rack, 1x15ml D17mm Rim, 2/pk
14.5
-0.3
30314852
Rack, 7x15ml D17mm RB, 2/pk
14.5
-0.3
30314853
Rack, 9x5-7ml D13mm RB, 2/pk
14.2
-0.6
30314854
Rack, 10x1.5ml D11mm, 2/pk
9.9
-4.9
30314855
Rack, 9x4.5ml D12mm FB, 2/pk
14.0
-0.8
30314856
Rack, 7x5-7ml D13mm RB, 2/pk
11.6
-3.2
30314857
Rack, 7x5-7ml D13mm Vac, 2/pk
14.2
-0.6
30314858
Rack, 5x4-10ml D16mm Sar, 2/pk
14.3
-0.5
30314823
Rotor, Sw ing out,
4x 100ml, ID,
Sealable
Blank
NO Bucket/Adapter
14.6
30314860
Bucket, 100ml w o Cap, Sealable,
2/pk
14.2
-0.4
30314861
Bucket, 100ml w/ Cap, Sealable, 2/pk
14.2
-0.4
30314864
Rack, 1x100ml D40mm RB, 2/pk
13.8
-0.8
30314865
Rack, 1x85ml D38mm RB, 2/pk
14.0
-0.6
30314866
Rack, 1x 50ml D34mm RB, 2/pk
14.0
-0.6
30314867
Rack, 1x50ml D29mm RB, 2/pk
14.0
-0.6
30314868
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.5
30314869
Rack, 1x50ml D29mm Rim, 2/pk
13.9
-0.7
30314870
Rack, 1x30ml D25mm Rim, 2/pk
13.1
-1.5
30314871
Rack, 1x25ml D25mm RB, 2/pk
14.0
-0.6
30314872
Rack, 1x15ml D17mm FA, 2/pk
14.1
-0.5
30314873
Rack, 2x15ml D17mm FA, 2/pk
14.1
-0.5
30314874
Rack, 2x15ml D17mm Rim, 2/pk
14.0
-0.6
30314875
Rack, 4x15ml D17mm RB, 2/pk
14.0
-0.6
30314876
Rack, 7x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314877
Rack, 5x1.5ml D11mm, 2/pk
14.0
-0.6
30314878
Rack, 3x10ml D17mm Sar, 2/pk
14.0
-0.6
30314879
Rack, 4x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314880
Rack, 4x10ml D17mm Vac, 2/pk
14.0
-0.6
30314881
Bucket, 2x50ml D29mm FA, w / Rack,
2/pk
14.6
0.0
30314882
Bucket, 3x15ml D17mm FA, w / Rack,
2/pk
14.6
0.0
30314883
Cy to, Rectangular Bucket, 2/pk
6.3
-8.3
30314824
Rotor, Sw ing out,
2x 3MTP, ID
Blank
NO Bucket/Adapter
12.0
30314891
Adapter, 48x 1.5/2.0ml D11mm, 2/pk
12.0
0.0
30314825
Rotor, Angle,
6x 85ml RB, ID, Hi
Blank
NO Bucket/Adapter
10.3
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
9.6
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
9.6
-0.7
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.0
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
9.8
-0.5
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
9.3
-1.0
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
10.3
0.0
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
9.5
-0.8
30314826
Rotor, Angle,
6x 85ml RB, ID
Blank
NO Bucket/Adapter
11.5
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.9
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
10.4
-1.1
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.9
-0.6
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
10.6
-0.9
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
10.4
-1.1
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
11.1
-0.4
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
10.4
-1.1
FR-44 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314827
Rotor, Angle,
4x 85ml RB, ID, Hi
Blank
NO Bucket/Adapter
9.2
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
8.5
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
8.4
-0.8
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
8.9
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
8.6
-0.6
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
8.3
-0.9
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
7.5
-1.7
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
8.3
-0.9
30314828
Rotor, Sw ing out,
16x 50ml, ID
Blank
NO Bucket/Adapter
16.5
30304375
Rack, 4x50ml D29mm FA, 2/pk
16.5
0.0
30314583
Rack, 10x15ml D17mm FA, 2/pk
16.5
0.0
30314829
Rotor, Angle,
10x 50ml FA, ID
Blank
NO Bucket/Adapter
13.0
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
12.8
-0.2
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.4
-2.6
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
8.9
-4.1
30314830
Rotor, Angle,
6x 50ml RB/FA, ID
Blank
NO Bucket/Adapter
11.0
30130891
Adapter, 1x 30ml D26mm, 2/pk
10.7
-0.3
30130892
Adapter, 1x 16ml D18mm, 2/pk
10.3
-0.7
30130893
Adapter, 1x 15ml D17mm, 2/pk
10.6
-0.4
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.4
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
8.3
-2.7
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
6.7
-4.3
30314831
Rotor, Angle,
6x 50ml RB, ID, Hi
Blank
NO Bucket/Adapter
8.4
30130891
Adapter, 1x 30ml D26mm, 2/pk
8.2
-0.2
30130892
Adapter, 1x 16ml D18mm, 2/pk
7.9
-0.5
30130893
Adapter, 1x 15ml D17mm, 2/pk
8.0
-0.4
30314892
Adapter, 1x 15ml D17.5mm FA, 2/pk
7.7
-0.7
30314832
Rotor, Angle,
30x 15ml RB/FA,
ID
Blank
NO Bucket/Adapter
12.5
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.3
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.5
-2.0
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
9.0
-3.5
30314834
Rotor, Angle,
12x 15ml RB/FA,
ID
Blank
NO Bucket/Adapter
11.0
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.6
-0.4
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
9.1
-1.9
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
7.7
-3.4
30314835
Rotor, Angle,
44x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.7
-0.7
30314836
Rotor, Angle,
30x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
9.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
9.1
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
8.4
-1.0
30314838
Rotor, Angle,
24x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314839
Rotor, Angle,
12x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
6.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
6.3
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
5.6
-0.9
Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-45
30314900
Adapter, 1x 1.5ml for ov er16000x g,
6/pk
6.4
-0.1
30553085
Rotor Sw ingOut
4x 750ml D99mm
ID SL
Blank
NO Bucket/Adapter
19.3
30553138
Adapter 32x 7ml D13mm RBx 2 V1
18.7
-0.6
30553135
Adapter 18x 15ml D17mm RBx 2 V1
18.8
-0.5
30553132
Adapter 8x 30ml D26mm RBx 2 V1
19.1
-0.2
30553131
Adapter 7x 50ml D29mm RBx 2 V1
19.1
-0.2
30553127
Adapter 3x 100ml D41mm RBx 2 V1
19.1
-0.2
30553125
Adapter 1x 250ml D62mm FBx 2 V1
18.9
-0.4
30553126
Adapter 1x 225/190ml D61mm FAx 2
V1
19.1
-0.2
30553128
Adapter 5x 50ml D29mm FAx 2 V1
19.1
-0.2
30553130
Adapter 7x 50ml D29mm FAx 2 V1
19.1
-0.2
30553133
Adapter 14x 15ml D17mm FAx 2 V1
19.2
-0.1
30553134
Adapter 17x 15ml D17mm FAx 2 V1
19.0
-0.3
30553140
Adapter 18x 5/10ml D16mm
Vac/Sarx 2 V1
19.0
-0.3
30559377
Adapter 38x 1.1/1.4ml D8mm Sarx2
V1
18.9
-0.4
30553136
Adapter 21x 10ml D16mm RBx 2 V1
18.9
-0.4
30553139
Adapter 27x 1.8/7ml D13mm
Vac/Sarx 2 V1
18.8
-0.5
30553086
Rotor Angle
4x 500ml D69mm
ID
Blank
NO Bucket/Adapter
14.5
30564850
Adapter 1x 250ml D61.5mm FBx 2 V1
13.7
-0.8
30559417
Adapter 1x 175/225ml D61.5mm
FAx 2 V1
13.4
-1.1
30559419
Adapter 1x 50ml D29.5mm FAx 2 V1
12.4
-2.1
30559416
Adapter 1x 250ml D60.5mm CBCx 2
V1
12.6
-1.9
30559421
Adapter 4x 30ml D25.5mm RBx 2 V1
13.8
-0.7
30559420
Adapter 3x 50ml D29mm RBx 2 V1
14.3
-0.2
30559422
Adapter 6x 15ml D17mm FAx 2 V1
13.8
-0.7
FR-46 Centrifugeuse Frontier
TM
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.7 Tableau 7 : Formulaire de rachat/Certificat de décontamination
Ce formulaire doit accompagner tous les retours dquipements et d'ensembles !
La déclaration complète sur la décontamination est un prequis avant tout traitement du produit retourné. Si aucune explicat ion correspondante
n'est incluse, il sera prodé à la décontamination aux frais du cl ient.
Prénom ; nom :_______________________________________
Socié/Entreprise :___________________________________
Ville : ______________________________________________
CODE POSTAL : _____________ lieu :____________________
Téléphone : _____________ Fax :____________________
E-Mail : _____________________________________
Pos. Crowd
Decontaminated
object
Serial number Description / Comment
1
2
3
4
Les pièces pertoriées ci-dessus sont-elles en contact avec les substances suivantes ?
Solutions aqueuses, tampons, acides, alcalis dangereux pour la santé :............... Oui Non
Agents infectieux potentiels : ................................................................................. Oui Non
Réactifs organiques et solvants : …………………………………………………….. Oui Non
Substances radioactives :............................................................. α.. β.. γ... Oui Non
Protéines dangereuses pour la santé :……………………………………………….. Oui Non
ADN : …………………………………………………………………………………..… Oui Non
Ces substances sont-elles entrées en contact avec lquipement/l'assemblage ? Oui Non
Laquelle, si oui :
________________________________________________________________
Description des mesures de décontamination des pièces répertoriées :
Je confirme que la décontamination est adéquate :
Entreprise/Service ________________________ Lieu et date : ________________________
Signature de la personne autorisée :_____________________________________
Compléter le formulaire
en lettres capitales
Skizzenkarte von FC5714 / FC5718 / FC5816 / FC5916
Abbildung.1
Skizzenkarte von FC5718R
Abbildung.2
Skizzenkarte von FC5816R / FC5916R
Abbildung.3
1 Zentrifugendeckel
2 Rotorfenster
3 Anzeige
4 Funktionstafel
5 Hauptstromschalter
6 Stromanschluss
7 Notfreigabe
4
1
7
4
2
1
3
5
6
5
6
7
2
3
4
1
5
6
7
2
3
Funktionsetikett für FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Abbildung.4
Abbildung.5
1. LCD-Anzeige
2. Programmeinstellungsmodell
3. Zentrifugation starten
4. RPM/RFC-Modell und Auswahl
5. Beschleunigungs-
/Verlangsamungsintensitätsmo
dell (Acc/Dec) und Auswahl
6. Temperatureinstellungsmodell
(Nur F5515R)
7. Zeiteinstellungsmodell
8. Speichereinstellungsmodell
9. Kurze/Schnelle
Drehzentrifugation
10. Freigabedeckel
11. Zentrifugation stoppen /
Einstellung
12. Anpassungsknopf/Wählscheibe:
Nummer ändern
LCD-Anzeige
Das folgende Bild zeigt die einzelnen Elemente der LCD-Anzeige.
Abbildung.6
1
2
3
4
5
6
7
8
9
10
11
12
Abbildung.7
Anzeigefelder:
A-1 Anzeigefeld – "RPM/RCF"
A-2 Anzeigefeld – "Acc/Dec" "Service"
A-3 Anzeigefeld – "Time/Prog"
A-4 Anzeigefeld – "Temp"
Mitteilungen/Logos der Anzeigefelder
M1 "close"
M2 "open"
M3 "rotor"
M4 "Rotor-Nr."
M5 "rpm"
M6 "rcf"
M7 "accel"
M8 "decel"
M9 "radius"
M10 "program"
M11 "error"
M12 "service"
M13 "h m s"
M14 "temperature"
M15 "precool"
M14
M5
M6
M4
A-1
A-2
A-3
A-4
M2
M3
M1
M7
M8
M12
M15
M9
M11
M10
M13
Rotor informations tabelle
Rotor
Nr.
Anzeige
Bestell-Nr.
Kapazitä t
Kompatibel
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
61
30304361
Rotor, Angle, 24x1.5/2.0ml, ID,
Sealable
20
30314820
Rotor, Swing out, 4x250ml, ID
21
30314821
Rotor, Angle, 6x250ml FB, ID
22
30314822
Rotor, Swing out, 4x100ml, ID
23
30314823
Rotor, Swing out, 4x100ml, ID,
Sealable
24
30314824
Rotor, Swing out, 2x3MTP, ID
25
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
26
30314826
Rotor, Angle, 6x85ml RB, ID
27
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
28
30314828
Rotor, Swing out, 16x50ml, ID
29
30314829
Rotor, Angle, 10x50ml FA, ID
30
30314830
Rotor, Angle, 6x50ml RB/FA, ID
31
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
32
30314832
Rotor, Angle, 30x15ml RB/FA, ID
33
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
34
30314834
Rotor, Angle, 12x15ml RB/FA, ID
35
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
36
30314836
Rotor, Angle, 30x1.5/2.0ml, ID,
Sealable
38
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
39
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
40
30314840
Rotor, Angle, 64x0.5ml, ID
41
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
85
30553085
Rotor SwingOut 4x750ml D99mm ID
SL
86
30553086
Rotor Angle 4x500ml D69mm ID
INHALTSANGABE
1. EINLEITUNG
..........................................................................................................................................................................1
1.1 BESCHREIBUNG UND VERWENDUNGSZWECK ....................................................................................................................1
1.2
KURZBESCHREIBUNG
...........................................................................................................................................................1
1.3 DEFINITION DER SIGNALWARNUNGEN UND SYMBOLE ......................................................................................................1
1.4
SICHERHEITSVORKEHRUNGE N
............................................................................................................................................2
1.4.1
BENUTZER
....................................................................................................................................................................2
1.4.2 ROTOR UND
ZUBEHÖ RTEILE
........................................................................................................................................2
1.4.3 MAßNAHMEN ZU IHREM EIGENEN
SCHUTZ
................................................................................................................2
1.4.4 SCHLIEN SIE DIE FOLGENDEN UMWELTEINFLÜSSE
AUS
..........................................................................................2
1.4.5 MAßNAHMEN R DIE BETRIEBSSICHERHEIT..............................................................................................................3
1.4.6 GEFAHR UND
VORKEHRUNGEN
...................................................................................................................................3
1.4.7 ABRZUNGEN, DIE IN DIESEM HANDBUCH BENUTZT WERDEN................................................................................3
2.
INSTALLATION
.......................................................................................................................................................................4
2.1 AUSPACKEN .........................................................................................................................................................................4
2.1.1 LIEFERPAKET ................................................................................................................................................................4
2.2 AUSWAHL DES STANDORTS.................................................................................................................................................4
2.3 INSTALLATION......................................................................................................................................................................4
2.4
SICHERHEITSVORKEHRUNGE N
HREND DES BETRIEBS UND
GARANTIE
.........................................................................5
3.
BETRIEB
.................................................................................................................................................................................5
3.1 BEFESTIGEN UND LADEN DES
ROTORS
................................................................................................................................5
3.1.1 INSTALLATION VON ROTOREN.....................................................................................................................................5
3.1.2 BELADEN VON WINKLIGEN ROTOREN .........................................................................................................................6
3.1.3 BELADEN VON
AUSSCHWENKROTORE N
......................................................................................................................6
3.1.4 BELADEN UND ÜBERLADEN DER
ROTOREN
.................................................................................................................7
3.1.5 ENTFERNUNG DES ROTORS .........................................................................................................................................7
3.2 KONTROLLE DES
DECKELS
....................................................................................................................................................7
3.2.1 DECKEL
Ö FFNEN
...........................................................................................................................................................7
3.2.2 DECKELVERSCHLUSS ....................................................................................................................................................8
3.3 VORAUSWAHL .....................................................................................................................................................................8
3.3.1 VORAUSWAHL DER GESCHWINDIGKEIT /
RCF-WERT
..................................................................................................8
3.3.2 VORAUSWAHL DER
LAUFZEIT
......................................................................................................................................9
3.3.3 VORAUSWAHL DER BREMSINTENSITÄT UND
BESCHLEUNIGUNG
...............................................................................9
3.3.4.VORAUSWAHL DER TEMPERATUR (NUR GEHLTE MODELLE)................................................................................10
3.3.5 VORKÜHLUNG (NUR GEKÜHLTE MODELLE)...............................................................................................................10
3.4
RADIUSKORREKTUR
...........................................................................................................................................................10
3.5 PROGRAMM ......................................................................................................................................................................11
3.5.1 SPEICHERUNG VON
PROGRAMMEN
..........................................................................................................................11
3.5.2 AUFRUF VON GESPEICHERTEN PROGRAMMEN ........................................................................................................12
3.5.3 VERLASSEN DES PROGRAMMMODUS .......................................................................................................................12
3.6 STARTEN UND STOPPEN DER ZENTRIFUGE .......................................................................................................................12
3.6.1 STARTEN DER
ZENTRIFUGE
........................................................................................................................................12
3.6.2 DIE
"STOP"-TASTE
......................................................................................................................................................13
3.7 FESTSTELLUNG EINES UNGLEICHGEWICHTS......................................................................................................................13
4.
EINSTELLUNG
......................................................................................................................................................................14
4.1
GRUNDANPASSUNGEN
......................................................................................................................................................14
4.1.1 ZUGRIFF AUF DEN MODUS
"BETRIEBSDATEN"
..........................................................................................................14
4.1.2
TEMPERATURANGABE
...............................................................................................................................................15
4.1.3 SIGNAL
EIN-/AUSSCHALTEN
.......................................................................................................................................15
4.1.4
LAUTSTÄRKE-VORAUSWAHL
DES
TONSIGNALS
.........................................................................................................16
4.1.5 LIEDAUSWAHL FÜR DAS TONSIGNAL AM ENDE DES DURCHLAUFS ..........................................................................16
4.1.6 TASTATURGERÄUSCH EIN-/
AUSSCHALTEN
...............................................................................................................16
4.1.7 BETRIEBSDATEN AUFRUFEN ......................................................................................................................................17
5.
WARTUNG...........................................................................................................................................................................18
5.1 WARTUNG UND REINIGUNG .............................................................................................................................................18
5.1.1 ALLGEMEIN ...............................................................................................................................................................18
5.1.2 REINIGUNG UND DESINFEKTION DER
EINHEIT
..........................................................................................................18
5.1.3 REINIGUNG UND DESINFEKTION DES
ROTORS
..........................................................................................................19
5.1.4 DESINFEKTION VON
ALUMINIUMROTOREN
..............................................................................................................19
5.1.5 DESINFEKTION DER
PP-ROTOREN
..............................................................................................................................19
5.1.6 GLASBRUCH ...............................................................................................................................................................19
5.2 LEBENSZEIT DER ROTOREN, BEHÄLTER,
ZUBEHÖRTEILE
...................................................................................................19
6.
PROBLEMBEHEBUNG
..........................................................................................................................................................20
6.1 FEHLERMITTEILUNG: URSACHE /
LÖ SUNG
........................................................................................................................20
6.2 UNTERSUCHUNG DER GLICHEN FEHLERMITTEILUNGEN UND IHRER LÖ SUNGEN.......................................................20
6.2.1 DECKELFREIGABE BEI STROMAUSFALL
(NOTFALLDECKELFREIGABE)
........................................................................20
6.2.2 BESCHREIBUNG DES
FEHLERMITTEILUNGSSYSTE MS
.................................................................................................21
7. ENTGEGENNAHME VON ZU REPARIERENDEN
ZENTRIFUGEN
............................................................................................21
8. TRANSPORT, LAGERUNG UND
ENTSORGUNG
....................................................................................................................22
8.1 TRANSPORT .......................................................................................................................................................................22
8.2
LAGERUNG
.........................................................................................................................................................................22
8.3 TRANSPORT, INSTALLATION, ÜBERTRAGUNG UND ENTSORGUNG DER ZENTRIFUGE
FC5714/FC5718/FC5718R/FC5816/FC5816R
..........................................................................................................................22
8.3.1 TRANSPORT................................................................................................................................................................22
8.3.2
INSTALLATION
............................................................................................................................................................22
8.3.3 VERPACKUNG.............................................................................................................................................................23
8.3.4 WEITERGABE DES
GE TS
.........................................................................................................................................23
9. TECHNISCHE DATEN............................................................................................................................................................24
9.1 SPEZIFIKATIONEN ..............................................................................................................................................................24
9.1.1 ZENTRIFUGE
FC5714
..................................................................................................................................................24
9.1.2 ZENTRIFUGE
FC5718
..................................................................................................................................................25
9.1.3 ZENTRIFUGE
FC5718R
................................................................................................................................................26
9.1.4 CENTRIFUGE
FC5816
..................................................................................................................................................27
9.1.5 ZENTRIFUGE
FC5816R
................................................................................................................................................28
9.2 ZEICHNUNGEN UND
AUSMAßE
.........................................................................................................................................29
10. COMPLIANCE ......................................................................................................................................................................31
11. ANHANG .............................................................................................................................................................................32
11.ϭ TABELLE ϭ: ZULÄSSIGES
NETTOGEWICHT
........................................................................................................................33
11.Ϯ TABELLE Ϯ: NIEDRIGSTE TEMPERATUREN BEI
MAX.-GESCHWINDIGKEIT
(NUR GEKÜHLTE MODELLE)...........................34
11.ϯ TABELLE ϯ: MAX. GESCHWINDIGKEIT UND RCF-WERTE R ZULÄSSIGE
ROTOREN
.......................................................͘35
11.ϰ TABELLE ϰ: BESCHLEUNIGUNGS- UND
VERLANGSAMUNGSZEIT E N
................................................................................37
11.ϱ TABELLE ϱ: FEHLERMITTEILUNGEN .................................................................................................................................40
11.ϲ TABELLE ϲ: RADIUSKORREKTUR UND ADAPTERSPEZIFIKATIONEN..................................................................................41
11.ϳ TABELLE ϳ:CKNAHMEFORMULAR / DEKONTAMINIERUNGSZERTIFIKAT....................................................................45
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-1
1. EINLEITUNG
1.2 Kurzbeschreibung
Die Modelle FC5714/FC5718/FC5816/FC5916 sind nicht gekühlte Universalzentrifugen.
Die Modelle FC5718R/FC5816R/FC5916R sind gekühlte Universalzentrifugen.
Alle drei Modelle werden in zwei Spannungsvarianten, 230V oder 120V angeboten.
Die Zentrifugen können mit Schwenk- und Winkelrotoren verwendet werden.
Alle Parameter sind über Tasten zugänglich und werden mit dem zentralen Einsteller ausgewählt. Alle
vorgewählten und aktuellen Werte werden auf der LCD-Anzeige dauerhaft angezeigt.
Die Zentrifuge wird durch einen wartungsfreien Induktionsmotor angetrieben.
Detaillierte technische Daten finden Sie im Abschnitt "Technische Daten".
1.3 Definition der Signalwarnungen und Symbole
Sicherheitsanmerkungen sind mit Signalwörtern und Warnsymbolen gekennzeichnet. Diese zeigen
Sicherheitsproblem und Warnungen an. Eine Missachtung der Sicherheitsanmerkungen kann zu Verletzungen an
Personen, Schäden an Instrumenten, sowie Fehlfunktionen und falschen Ergebnissen führen.
Der Grad der Gefahr ist ein Teil einer Sicherheitsanmerkung und unterscheidet die möglichen Ergebnisse ihrer
Nichteinhaltung.
Signalwörter
GEFAHR
Wird zu schweren Verletzungen oder zum Tode führen, falls nicht vermieden.
WARNUNG
Für eine gefährliche Situation mit mittlerem Risiko, die möglicherweise zu
Verletzungen oder zum Tode führt, falls nicht vermieden.
VORSICHT
Für eine gefährliche Situation mit geringem Risiko, die du eine Schaden am Gerät
oder
Eigentum oder zu einem Datenverlust oder zu Verletzungen führt, falls nicht
vermieden.
ACHTUNG
Für wichtige Informationen über das Produkt Kann zu Schäden an der Ausrüstung führen,
falls nicht vermieden
ANMERKUNG
Für nützliche Informationen über das Produkt
Warnsymbole
Allgemeine Gefahr Stromschlaggefahr
Wechselstrom Biologische Gefahr
Explosion Zerkleinerung
1.1 Beschreibung und Verwendungszweck
Vielen Dank dafür, dass Sie ein OHAUS-Produkt ausgewählt haben.
Alle Symbole stellen Sicherheitsanweisungen dar und weisen auf potentielle Gefahrensituationen hin. Bitte lesen
Sie das Handbuch vollständig durch, bevor Sie die Frontier
TM
MultiPro-Zentrifugen verwenden, um
Fehlbedienungen zu vermeiden.
Frontier
TM
MultiPro Zentrifugen wurden für die Trennung von Materialien oder Gemischen mit unterschiedlichen
Dichten entwickelt.
OHAUS-Zentrifugen sind ausschlilich für die Innenanwendung und für die Anwendung durch qualifiziertes
Personal vorgesehen.
DE-2 Frontier
TM
Zentrifuge
Warn- und Informationsschilder auf der Zentrifugenoberfläche
Es müssen stets vier Trägers an vier Stellen bei Ausschwenkrotoren verwendet
werden, da ansonsten Beschädigungen an der Zentrifuge auftreten können.
Eine derartige Beschädigung wird von der Produktgarantie nicht abgedeckt.
Achtung! Prüfen Sie die Befestigung der Rotormutter vor jedem Lauf.
Ziehen Sie den Netzstecker, bevor Sie das Gehäuse oder die Notfreigabe öffnen.
Drehrichtung - Drehung im Uhrzeigersinn für den Rotorantrieb
Referenz zum Laden von Rotoren
1.4 Sicherheitsvorkehrungen
1.4.1 Benutzer
OHAUS-Zentrifugen sind ausschließlich für die Innenanwendung und für die Anwendung durch qualifiziertes
Personal vorgesehen. Das Gerät kann nur von einem geschulten Fachmitarbeiter betrieben werden. Sie müssen
das Betriebshandbuch gründlich durchlesen und sich mit der Funktion des Gerätes vertraut machen.
1.4.2 Rotor und Zubehörteile
Es sollten nur Originalrotoren und Zubehörteile von OHAUS verwendet werden. Jede andere Nutzung oder
beabsichtigte Nutzung wird als ungeeignet erachtet. OHAUS haftet nicht für Schäden, die aus einer ungeeigneten
Anwendung resultieren.
VORSICHT:
Lesen Sie alle Sicherheitswarnungen durch, bevor Sie diese Ausrüstung installieren, anschließen oder warten.
Ein Versäumnis, diese Warnungen zu befolgen, könnte zu Verletzungen an Personen und/oder
Beschädigungen am Eigentum führen. Bewahren Sie die Anweisungen zwecks zukünftiger Bezugnahme auf.
1.4.3 Maßnahmen zu Ihrem eigenen Schutz
WARNUNG: Arbeiten Sie niemals in einer explosionsgefährdeten Umgebung! Das Gehäuse des
Instruments ist nicht gasdicht. (Explosionsgefahr aufgrund von Funkenbildung, Korrosion hervorgerufen
durch das Eindringen von Gasen)
WARNUNG: Befolgen Sie, wenn Sie Chemikalien und Lösungsmittel benutzen, die Anweisungen des
Herstellers und die allgemeinen Regeln zur Laborsicherheit.
WARNUNG: Die Zentrifuge ist nicht versiegelt. Benutzen Sie geeignete Schutzmaßnahmen, wenn Sie
die Zentrifuge für infektiöse und pathogene Proben. Befolgen Sie angemessene Sicherheitsmaßnahmen,
wenn Sie mit diesen Proben arbeiten.
1.4.4 Schließen Sie die folgenden Umwelteinflüsse aus
Starke Vibrationen
Direktes Sonnenlicht
Atmosphärische Feuchtigkeit grösser als 80%
Korrosive Gase vorhanden
Temperaturen unter 2 °C und über 35 °C
Starke elektrische oder magnetische Felder:
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Frontier
TM
Zentrifuge DE-3
WARNUNG:
Innerhalb des Gehäuses besteht die Gefahr von elektrischen Stromschlägen. Das Gehäuse sollten nur
von autorisiertem und qualifiziertem Personal geöffnet werden. Entfernen Sie vor dem Öffnen alle
Stromverbindungen zur Einheit.
1.4.5 Maßnahmen für die Betriebssicherheit
Entschrauben Sie nicht die beiden Hälften des Gehäuses.
Trocknen Sie jede ausgelaufene Flüssigkeit sofort ab! Das Instrument ist nicht wasserdicht.
Prüfen Sie, dass die Eingangsspannung und die Steckerart des Geräts mit der lokalen Stromversorgung
vereinbar sind.
Stecken Sie das Stromkabel nur in einen ordnungsgemäß geerdeten Stromempfänger.
Benutzen Sie nur ein Stromkabel mit einem Nennwert, der die Spezifikationen auf dem Ausrüstungsetikett überschreitet.
Platzieren Sie die Ausrüstung nicht so, dass es schwierig ist, das Stromkabel vom Stromempfänger zu trennen.
Stellen Sie sicher, dass das Stromkabel kein potentielles Hindernis oder eine Stolperfalle darstellt.
Die Ausrüstung ist nur für den Innengebrauch vorgesehen. Benutzen Sie die Ausrüstung nur in trockenen Umgebungen.
Benutzen Sie nur angemessene Zubehörteile.
Betreiben Sie die Ausrüstung nur unter Umgebungsbedingungen, die in diesen Anweisungen festgelegt werden.
Trennen Sie die Ausrüstung bei der Reinigung von der Stromversorgung.
Betreiben Sie die Ausrüstung nicht in gefährlichen oder unstabilen Umgebungen.
Instandhaltungsarbeiten sollten nur von autorisiertem Personal durchgeführt werden.
1.4.6 Gefahr und Vorkehrungen
Um Mensch und Umwelt zu schützen, sollten die folgenden Vorkehrungen eingehalten werden:
Während der Zentrifugation ist der Aufenthalt von Personen und die Aufstellung von gefährlichen Materialien
innerhalb von 30 cm um die Zentrifuge herum, gemäß den Bestimmungen von EN 61010-2-020. verboten.
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R sind nicht explosionssicher und darf daher nicht in
expliosionsgefährdeten Bereichen oder Standorten betrieben werden. Die Zentrifugation von brennbaren,
explosiven, radioaktiven oder solchen Substanzen, die chemisch mit hoher Energie reagieren, ist strengstens
verboten. Die endgültige Entscheidung in Bezug auf die Risiken in Verbindung mit der Verwendung von solchen
Substanzen liegt in der Verantwortung der Benutzer der Zentrifuge.
Drehen Sie niemals toxisches oder pathogenes Material ohne angemessene Sicherheitsvorkehrungen, d.h.
Zentrifugation von Kübeln, / Rohren mit fehlender oder beschädigter hermetischer Versiegelung ist strengstens
verboten. Der Benutzer ist verpflichtet, ein angemessenes Desinfektionsverfahren durchzuführen, für den Fall,
dass gefährliche Substanzen die Zentrifuge und/oder ihre Zubehörteile kontaminiert haben. Wenn infektiöse
Substanzen zentrifugiert werden, achten Sie stets auf die allgemeinen Laborvorkehrungen. Falls erforderlich,
setzen Sie sich mit Ihrem Sicherheitsbeauftragten in Verbindung!
Es ist verboten, die Zentrifuge mit Rotoren zu betreiben, die nicht für diese Einheit aufgeführt sind.
Öffnen Sie den Deckel der Zentrifuge unter keinen Umständen, während der Rotor noch läuft oder sich mit einer
Geschwindigkeit von> 2m / s dreht
1.4.7 Abkürzungen, die in diesem Handbuch benutzt werden
Symbol/Abkürzungen
Einheit
Beschreibung
RPM
[min
-1
]
rpm
Umdrehungen pro Minute
RCF
[x g]
relative Zentrifugalkraft
PCR
Polymerasekettenreaktion
PP
-
Polypropylen
PC
-
Polykarbonat
accel
-
Beschleunigung
decel
-
Verlangsamung
prog
-
Programm
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-4 Frontier
TM
Zentrifuge
2. INSTALLATION
2.1 Auspacken
Entnehmen Sie Ihre Zentrifuge und jede ihrer Komponenten vorsichtig aus der Verpackung. Die enthaltenen
Komponenten variieren je nach Zentrifugenmodell (siehe Tabelle unten). Sichern Sie die Verpackung, um eine
sichere Lagerung und Transport zu gewährleisten. Die Bedienungsanleitung muss immer mit der Zentrifuge
aufbewahrt werden!
Rotor(en) / Zubehörteile werden separat verpackt.
WARNUNG: Hebegefahr. Das Heben durch eine einzige Person kann Unfälle verursachen.
Benutzen Sie ein mechanisches Hebegerät oder verknüpfen Sie Hebeverfahren, wenn Sie die
Ausrüstung heben oder bewegen.
Siehe Abschnitt 8.3 für Details über das Anheben des Gerätes
2.1.1 Lieferpaket
2.2 Auswahl des Standorts
ANMERKUNG
Vermeiden Sie übermäßige Vibrationen, Wärmequellen, Luftstrom oder schnelle
Temperaturveränderungen.
Die Zentrifuge sollte auf einer ebenen und festen Oberfläche installiert werden, möglichst auf einem
Laborschrank / Tisch oder einer anderen festen, vibrationsfreien Oberfläche.
Während der Zentrifugation muss die Zentrifuge so platziert werden, dass auf jeder Seite des Gerätes ein
Mindestabstand von 30 cm entsprechend den Normen EN 61010-2-020 vorhanden ist.
Stellen Sie die Zentrifuge nicht neben ein Fenster oder eine Heizung, wo sie einer übermäßigen Hitze
ausgesetzt sein könnte, da die Leistung des Gerätes auf einer Umgebungstemperatur von 23 ° C basiert.
2.3 Installation
Befolgen Sie die folgenden Schritte:
Prüfen Sie, ob die Stromversorgung mit den auf dem Typenschild des Herstellers angegebenen Werten
übereinstimmt.
Für FC5714/FC5718/FC581 sollte die Stromleitung durch einen Leistungsschalter von 10 A (Typ K) geschützt
werden.
Im Notfall muss ein Notausschalter außerhalb des Raumes installiert werden, um die Stromversorgung vom
Gerät zu trennen.
Verbinden Sie die Zentrifuge mit einer geerdeten Steckdose.
Schließen Sie die Zentrifuge ans Netz an. (Die Steckdose für das Stromkabel muss für die Trennung leicht
zugänglich sein)
Schalten Sie das Instrument ein, indem Sie den Hauptstromschalter benutzen.
Öffnen Sie den Deckel, indem Sie die Tür öffnen-Taste verwenden.
Entfernen Sie das Transportsicherung vom Motor.
Menge
Beschreib ung
1
Zentrifuge
FC5714/F C5718/F C5718R/FC5816/F C5816R
/FC5916/FC5916R
1
Stromkabel
1
G ar anti ek ar te
1
Betriebshandbuch/Kurzanleitung
1
Rotors c hl üss el
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
r FC5718R/FC5816R /FC5916/FC5916R sollte die Stromleitung durch einen Leistungsschalter von 16 A
(Typ K) geschützt werden.
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-5
2.4 Sicherheitsvorkehrungen während des Betriebs und Garantie
Betreiben Sie die Zentrifuge nicht, wenn sie nicht korrekt installiert ist.
Lehnen Sie sich während des Betriebs nicht an die Zentrifuge.
Halten Sie aus betrieblichen Gründen einen Abstand von mehr als 30 cm.
Stellen Sie keine potenziell gefährlichen Stoffe innerhalb des Sicherheitsbereichs von 30 cm.
Betreiben Sie die Zentrifuge nicht, wenn sie zerlegt wurde (z. B. ohne Gehäuse).
Betreiben Sie die Zentrifuge nicht laufen lassen, wenn an mechanischen oder elektrischen Bauteile gearbeitet
wird.
Verwenden Sie keine Zubehörteile wie Rotoren und Behälter, die nicht exklusiv von der OHAUS Corporation
zugelassen werden, außer handelsüblichen Zentrifugenröhrchen aus Glas oder Kunststoff.
Drehen Sie keine extrem korrosiven Stoffe, da diese die Materialien beschädigen oder schwächen können.
Die Zentrifuge darf nicht mit Rotoren oder Behälter betrieben werden, die Anzeichen auf Korrosion oder
mechanischen Beschädigungen aufweisen.
Der Hersteller ist für die Sicherheit und Zuverlässigkeit der Zentrifuge verantwortlich, nur wenn:
1) Das Gerät wird gemäß dieser Bedienungsanleitung betrieben.
2) Änderungen, Reparaturen oder sonstige Einstellungen werden von durch OHAUS autorisiertem
Personal durchgeführt und die Elektroinstallation entspricht dem entsprechenden Electrical Code.
ANMERKUNG
Garantie
Die Zentrifuge wurde einer gründlichen Prüfung und Qualitätskontrolle unterzogen. Im unwahrscheinlichen Fall von
Fertigungsfehlern, werden die Zentrifugen und Rotoren durch eine Garantie abgedeckt. Diese Garantie erlischt bei
unsachgemäßer Handlung, Beschädigung und Nachlässigkeit sowie bei unsachgemäßen Ersatzteilen und / oder
Zubehör oder unberechtigter Modifikation des Gerätes.
Der Hersteller behält sich die technischen Modifikationsrechte hinsichtlich technischer Verbesserungen vor!
3. BETRIEB
3.1 Befestigen und Laden des Rotors
3.1.1 Installation von Rotoren
Reinigen Sie die Antriebswelle sowie die Spannzange mit einem sauberen, fettfreien Tuch. Setzen Sie den Rotor
auf die Antriebswelle. (Siehe Abbildung unten). Achten Sie darauf, dass der Rotor vollständig auf der Motorwelle
montiert ist.
Abbildung.8
Abbildung. 9 Abbildung.10
Motorwelle und Kammer
Mutter für
Rotor
Werkzeug für
Rotor mit Mutter
Werkzeug für
Rotor ohne
Mutter
Schnappdeckel
Schraubdeckel
DE-6 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Halten Sie den Rotor mit einer Hand fest und befestigen Sie den Rotor an der Welle, indem Sie die
Befestigungsmutter im Uhrzeigersinn drehen. Ziehen Sie die Befestigungsmutter mit dem beigefügten
Rotorschlüssel fest (Siehe Abbildungen 9-10)
Wir werden ein Werkzeug für einen mutterlosen Rotor mit Zentrifuge bereitstellen, das Werkzeug für den
Mutterrotor wird mit Rotor bereitgestellt.
ACHTUNG!
Vergewissern Sie sich, dass die Befestigungsschraube vor jedem Lauf ordnungsgemäß installiert ist.
(Siehe Abbildung 9-10)
Die Zentrifuge darf nicht mit Rotoren oder Behälter betrieben werden, die Anzeichen auf Korrosion
oder mechanischen Beschädigungen aufweisen.
Betreiben Sie die Zentrifuge nicht mit extrem korrosiven Substanzen, die den Rotor, die Behälter und
Materialien beschädigen könnten.
Bei Fragen wenden Sie sich bitte an den Hersteller!
3.1.2 Beladen von winkligen Rotoren
Rotoren müssen symmetrisch und mit gleichem Gewicht geladen werden (Siehe Abbildung unten). Dieser Adapter
kann nur mit den entsprechenden Gefäßen beladen werden. Die Gewichtsdifferenzen zwischen gefüllten Gefäßen
sollten so gering wie möglich gehalten werden. Daher empfehlen wir, die mit einer Waage auszutarieren. Dies
reduziert die Abnutzung des Antriebs und den akustischen Betriebslärm
Die Maximallast pro Loch wird auf dem Rotor angegeben.
Abbildung.11 FALSCH Abbildung.12 KORREKT ( 6 .Röhren)
3.1.3 Beladen von Ausschwenkrotoren
Beladen der Behälter / Gefäßen muss in Übereinstimmung mit der unteren Abbildung erfolgen.
Sie darf z.B. von einem 4-Place-Rotor mit nur 2 beladenen Behälter betrieben werden. Doch die beladenen
Behälter müssen sich gegenüberstehen. Stellen Sie sicher, dass auch die unbeladenen Behälter in den Rotor
eingelegt werden (siehe unten)
Im Prinzip können Ausschwenkrotoren nicht in Betrieb genommen werden, bis nicht alle Behälter oder Gestelle in
den Rotor gestellt sind.
Die Schrauben am Rotor müssen mit dem "High TEF Öl" gefettet werden. Die Probenröhrchen müssen nach
Augenmaß gleichmäßig gefüllt und in die Bohrungen oder Rohrständer gesteckt werden. Die Gewichtsdifferenz der
geladenen Behälter sollte 1,0 g nicht überschreiten.
ACHTUNG!
Ausschwenkrotoren können nur in Betrieb genommen werden, wenn alle Standorte mit vier
Behälter oder vier Trägern gefüllt sind - Mischen Sie keine Behälter und Träger zusammen!
ACHTUNG!
Die Zentrifuge darf nicht mit Rotoren oder Behälter betrieben werden, die Anzeichen auf Korrosion
oder mechanischen Beschädigungen aufweisen.
Betreiben Sie die Zentrifuge nicht mit extrem korrosiven Substanzen, die den Rotor und die
Behälter beschädigen könnten.
Bei Fragen wenden Sie sich bitte an den Hersteller!
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-7
Abbildung.13 FALSCH Abbildung.14 KORREKT
3.1.4 Beladen und überladen der Rotoren
Alle zugelassenen Rotoren werden mit ihrer Höchstgeschwindigkeit und dem maximalen Füllgewicht in "Tabelle 2
Zulässiges Nettogewicht" (siehe ANHANG) aufgeführt.
Die für einen Rotor zulässige Höchstbelastung, die vom Hersteller bestimmt wird, sowie die Höchstgeschwindigkeit
für diesen Rotor (siehe Etikett am Rotor) darf nicht überschritten werden. Die Flüssigkeiten, mit denen die Rotoren
beladen werden, sollten eine maximale homogene Dichte von 1,2 g/ml oder weniger haben, wenn der Rotor mit
maximaler Geschwindigkeit läuft.
Um Flüssigkeiten mit einer höheren Dichte zu drehen, muss die Geschwindigkeit nach folgender Formel reduziert
werden:
Reduzierte Geschwindigkeit n
red
=
1,2
𝑖𝑔𝑒𝑟 𝑑𝑒𝑛𝑠𝑖𝑡𝑦
x max. Geschwindigkeit (n
max
) des Rotors
Beispiel:
n
red
=
1,2
1,7
x 4.000 = 3.360 rpm
Bei Fragen wenden Sie sich bitte an den Hersteller!
Um die relative Zentrifugalkraft (RCF/G-Kraft) für einen bestimmten Adapter zu bestimmen, können Sie sie nach
DIN 58 970 mit der beigefügten Formel berechnen:
RCF = 1.117862*10
5
*n
2
*r
max
n: Umdrehungen pro Minute (RPM)
r
max
: max. Zentrifugationsradius in cm bei Verwendung derhrenböden
3.1.5 Entfernung des Rotors
Lösen Sie die Rotor-Befestigungsmutter vollständig (über den Steifpunkt schrauben) und heben Sie den Rotor
vertikal aus der Zentrifuge heraus. (Siehe Abbildung 9 und 10).
3.2 Kontrolle des Deckels
3.2.1 Deckel öffnen
Nach dem Lauf erscheint, wenn der Deckel der Zentrifuge offen ist, das Wort "schließen" (M1) auf der Anzeige
"RPM | RCF" (A-1). Zusätzlich, wenn sich ein Rotor in der Zentrifuge befindet, erscheint das Wort "Rotor"(M3) wie
auch die Code-Nummer des entsprechenden Rotors, der sich im Zentrifugensystem befindet "71" (M4). Wenn
sich kein Rotor in der Zentrifuge befindet, blinkt das Wort "Rotor" (M3) und es erscheint zusätzlich das Wort"kein"
(M4). Durch Drücken der Taste "Tür öffnen" (7) können Sie den Deckel der Zentrifuge freigeben. Sobald der
elektromagnetische Deckel vollkommen freigegeben ist, erscheint das Wort "offen" (M2). Nun können Sie den
Deckel der Zentrifuge öffnen.
Bitte siehe Abbildung 15 unten zwecks Referenz
DE-8 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Abbildung. 15
Während des Betriebs können Sie den Rotortyp zu jeder Zeit durch Drücken der Taste "Tür öffnen" (7) aufrufen.
3.2.2 Deckelverschluss
Der Deckel sollte nur leicht herabgedrückt werden. Ein elektromagnetischer Deckelverschluss schließt den Deckel,
gleichzeitig verschwindet das Wort "offen" (M2) (siehe Abbildung 15).
Als ein Zeichen dafür, dass die Zentrifuge Startbereit ist, erscheint in der Anzeige "RPM | RCF" (A-1) das Wort
"schließen" (M1). Gleichzeitig wird auch das Wort "Rotor" (M3) angezeigt, wie auch die Codenummer des Rotors,
die sich im Zentrifugensystem, "Nr 71" (M4) befindet. Damit werden alle rotorspezifischen Daten, wie z.B.
maximale Geschwindigkeit, Beschleunigung usw. übernommen
ACHTUNG:
Greifen Sie mit Ihren Fingern nicht zwischen den Deckel und das Gerät oder den Sperrmechanismus,
wenn Sie den Deckel verschließen!
3.3 Vorauswahl
3.3.1 Vorauswahl der Geschwindigkeit / RCF-Wert
Diese Vorauswahl wird durch die Taste "RPM | RCF" (4) aktiviert (siehe Abbildung 16 unten). Wenn Sie die Taste
einmal drücken, blinkt das Wort "rpm" (M5). Wenn Sie die Taste zweimal drücken, kann die Vorauswahl der
Zentrifugalkräfte ausgewählt werden. Dann erscheint das blinkende Wort "rcf" (M6). Sie können die gewünschten
Werte mit dem Anpassungsknopf (1) einstellen. In der Anzeige (A-1) wird stets der regulierte Wert angezeigt, bevor,
während und nach dem Betrieb.
Abbildung 16
Solange kein Rotor eingesetzt ist, kann die Geschwindigkeit zwischen 200 rpm und der maximalen
Umdrehungszahl der Zentrifuge eingestellt werden.
Wenn ein Rotor in der Zentrifuge vorhanden ist, kann die Geschwindigkeit nur bis zur maximal zulässigen
Umdrehungszahl dieses Rotors vorgewählt werden. Genauso verhält es sich mit der Vorauswahl des RCF-Wertes.
M5
A-1
4
5
M6
M3
A-1
M2 M4 M1
7
1
1
Der Einstellbereich liegt zwischen der Fliehkraft bei 200 U / min und der maximal zulässigen Fliehkraft des Rotors.
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-9
Siehe "Tabelle 3: Max. Geschwindigkeit und RCF-Werte für zulässigen Rotor" (Siehe ANHANG). Dort sind
alle wichtigen Werte aufgeführt.
ACHTUNG:
Bitte prüfen Sie auch die maximal zulässigen Drehzahlen Ihrer Teströhrchen mit dem Hersteller.
3.3.2 Vorauswahl der Laufzeit
Die Laufzeit kann in drei verschiedenen Reichweiten von 10 Sekunden bis zu 99 Stunden 59 Minuten
vorausgewählt werden.
1. Reichweite von 10 Sekunden bis zu 59 Minuten 50 Sekunden in Schritten von 10 Sekunden
2. Reichweite von 1 Stunde bis zu 99 Stunden 59 Minuten in Schritten von 1 Minute.
3. Der dauerhafte Betrieb "cont", der durch die Taste "Stop"(10) unterbrochen werden kann (siehe Abbildung 17).
Die Betriebszeit kann vorab ausgewählt werden, während der Deckel offen oder geschlossen ist.
Um die Einstellung der Betriebszeit einzustellen, drücken Sie die Taste "time" (6).
Auf der Anzeige "Time/Prog" (A-3) blinkt die Anzeige "m : s" oder "h : m", je nach vorheriger Einstellung.
Um den gewünschten Wert einzustellen, benutzen Sie den Anpassknopf (1). Nachdem 59 min 50 sec überschritten
werden, wechselt die Anzeige automatisch auf "h : m". Nachzem 99 hours 59 min überschritten werden, erscheint
das Wort "cont" in der Anzeige "Time/Prog" (A-3). Dieser dauerhafte Betrieb kann nur durch Drücken der Taste
"Stop" (10) unterbrochen werden. Der Countdown beginnt so bald die eingestellte Geschwindigkeit erreicht ist.
Die Anzeige zeigt immer die verbliebene Zeit. (Siehe Abbildung 17)
Abbildung 17
3.3.3 Vorauswahl der Bremsintensität und Beschleunigung
Diese Funktion wird durch die Taste "Accel/Decel" (5) aktiviert (siehe Abbildung18).
Wenn die Taste nur einmal gedrückt wird, blinkt das Wort "accel" (M7) in der Anzeige "Acc/Dec" (A-2). Die
gewünschte Beschleunigung kann vorab durch den Anpassungsknopf (1) ausgewählt werden. Der Wert 0
entspricht der geringsten und der Wert 9 der höchsten Beschleunigung.
Durch zweimaliges Drücken der Taste "Accel/Decel" (5) zeigt die Anzeige "Acc/Dec" (A-2) das Wort "decel"(M8)
an. Nun kann die gewünschte Bremsintensität vorab durch den Anpassungsknopf (1) ausgewaehlt werden. Der
Wert 9 entspricht der kürzesten und der Wert 0 der längsten möglichen Bremszeit.
Siehe "Tabelle 5: Beschleunigungs- und Verlangsamungszeiten" (ANHANG). Dort werden die
Beschleunigungs- und Verlangsamungszeiten für die Beschleunigungs- und Verzögerungsphasen 0-9 für zulässige
Rotoren gezeigt.
Abbildung. 18
6
A-3
1
10
M7
M8
5
A-2
1
DE-10 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
3.3.4.Vorauswahl der Temperatur (Nur gekühlte Modelle)
Diese Funktion wird von der Taste "Temp/Setup" (13) aktiviert. Nach dem Drücken dieser Taste blinkt in der
Anzeige "Time/Prog" die Anzeige "°C" (A-4). (1) die gewünschte Testtemperatur kann in Schritten von 1°C in
einer Reichweite von -20°C bis zu +40°C vorab ausgewählt werden.
Der Wert wird ständig auf der Anzeige angegeben (Abbildung 19) - vor, während und nach dem Betrieb.
Bitte beachten Sie die entsprechenden, niedrigsten Temperaturen der Rotoren bei maximaler Geschwindigkeit!
Abbildung 19
3.3.5 Vorkühlung (nur gekühlte Modelle)
Wenn die Proben temparaturempfindlich sind, ist es sinnvoll, die Zentrifuge, den Rotor und möglicherweise die
Behälter auf die gewünschte Arbeitstemperatur vorzukühlen. Fügen Sie daher den gewünschten Rotor ein und
stellen Sie die entsprechende Temperatur vorab ein. Durch gleichzeitiges Drücken der Tasten "Temp/Setup" (13)
(siehe Abbildung 20) und "Time" (6) starten Sie den Durchlauf. Während des Betriebs waehlt die Einheit
automatisch eine Rotationsgeschwindigkeit, die 20% der zulässigen Rotationsgeschwindigkeit des entsprechenden
Rotors entspricht. Nachdem die Vorstarttemperatur erreicht ist, können Sie den Vorkühldurchlauf mit der "Stop"-
Taste (10) verlassen.
Je nach eingefügtem Rotor, dauert die Vorkühlung ungefähr zwischen 10 und 20 min.
Abbildung. 20
3.4 Radiuskorrektur
Wenn Sie Adapter oder Reduzierer benutzen, könnte es den Zentrifugalradius des entsprechenden Rotors
verändern. In diesem Fall können Sie den korrekten Radius manuell korrigieren. Gehen Sie bitte wie folgt vor:
Schliessen Sie den Deckel, drücken Sie dann die Taste "Time" (6) (siehe Abbildung 21) und die Taste
"Prog/Setup" (11) zur gleichen Zeit und halten Sie sie.
Auf der Anzeige "Time/Prog" (A-3) erscheint das Wort "radius" (M9). Durch den Anpassungsknopf (1) können
Sie die entsprechende Radiuskorrektur in Schritten von 0,1 cm (Siehe Tabelle 7, ANHANG) vorab auswählen.
Sobald Sie die Radiuskorrektur eingestellt haben, erscheint das Wort "Radius" (M9). Das Wort wird sichtbar sein,
bis Sie die Radiuskorrektur auf 0 zurücksetzen.
1
13
6
10
A-4
A-4
13
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-11
Abbildung. 21.
3.5 Programm
3.5.1 Speicherung von Programmen
Sie können bis zu 99 Durchläufe mit allen relevanten Parametern speichern, einschließlich der benutzten Rotoren.
Sie können jede freie Programmnummer benutzen, und sie erneut aufrufen.
Setzen Sie den erforderlichen Rotor in die Zentrifuge. Durch Drücken der Taste "Prog/Setup" (11) erscheint auf
der Anzeige "Time/Prog" (A-3) das Wort "programm". Mit dem Anpassungsknopf (1) können Sie die gewünschte
Programmnummer auswählen.
Wenn eine Programmnummer bereits belegt ist, erscheinen auf der Anzeige "RPM | RCF" (A-1) die Worte "Rotor"
(M3) und "xx" (M4). Im Falle einer freien Programmnummer erscheint 0
Abbildung. 22
Schließen Sie den Deckel der Zentrifuge. Fahren Sie nun fort wie zuvor beschrieben, um alle wichtigen
Durchlaufparameter einzustellen. Wenn der Deckel nicht geschlossen ist, wenn Sie das Programm speichern,
blinken auf der Anzeige "RPM/RCF" (A-1) die Worte "FirSt" und "CLOSE Lid" (siehe Abbildung 23) abwechselnd.
Wenn Sie den Durchlauf starten möchten, ohne das Programm zu speichern, blinken auf der Anzeige "RPM/RCF"
(A-1) die Worte "First" und "PrESS StoreE" (siehe Abbildung 24) abwechselnd.
Abbildung 23 Abbildung. 24
11
A-3
A-1
M3
M4
M10
1
12
1
11
A-3
M9
6
DE-12 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Zur Anpassung der Daten drücken Sie die Taste "Store" (12) (siehe Abbildungen 23 und 24) für ungefähr 1
Sekunde. Wenn das Programm korrekt gespeichert wird, erscheint das Wort "StorE" auf der Anzeige "RPM/RCF"
(A-1). Als ein Ergebnis verschwindet das Wort "programm" (M10).
Sobald die Taste "Store" (12) freigegeben wird, erscheint das Wort "programm xx" (M10) erneut – das (xx) steht
für den ausgewählten Programmstandort.
Wenn die Programmnummern belegt sind, können Sie eine alte Nummer nehmen, die nicht mehr benötigt wird und
nur die neuen Parameter eingeben.
3.5.2 Aufruf von gespeicherten Programmen
Um gespeicherte Programme aufzurufen, drücken Sie die Taste "Prog/Setup" (11) (siehe Abbildung 25), während
der Deckel bereits geschlossen ist. Innerhalb der Anzeige "Time/Prog" (A-3) erscheint "programm –"(M10). Die
gewünschte Programmnummer kann vorab mit dem Anpassungsknopf (1) ausgewählt werden.
Auf den entsprechenden Anzeigen erscheinen die gespeicherten Werte für dieses Programm.
Wenn sich der falsche Rotor für das vorab ausgewählte Programm innerhalb der Zentrifuge befindet blinkt auf der
Anzeige "RPM | RCF" (A-1) das Wort "rotor" (M3). Gleichzeitig werden das Wort "FALSE" und die gespeicherte
Rotornummer " xx" (M4) abwechselnd blinken.
Abbildung. 25
3.5.3 Verlassen des Programmmodus
Um den Programmmodus zu verlassen, drücken Sie lediglich die Taste "Prog/Setup" (11) (siehe Abbildung 25).
Dann erscheint innerhalb der Anzeige "Time/Prog" das Wort "programm" .
Stellen Sie die Anzeige auf "programm--" (M10), indem Sie den Anpassungsknopf benutzen (1).
3.6 Starten und Stoppen der Zentrifuge
3.6.1 Starten der Zentrifuge
Sie können die Zentrifuge entweder mit der Taste"Start" (9) (siehe Abbildung 26) oder der Taste "Quick Spin" (8)
starten.
Mit der Taste "Start" (9) können Sie gespeicherte Durchläufe oder Durchläufe mit manuell vorab ausgewählten
Parametern starten.
Wenn die entsprechende, vorab ausgewählte Laufzeit abgelaufen ist, wird die Zentrifuge automatisch anhalten.
Mit der Taste "Quick Spin" (8) können Sie Durchläufe starten, die nur einige Sekunden dauern werden.
Durch Drüken der Taste "Quick Spin" (8) beschleunigt die Zentrifuge auf die vorab ausgewählte Umdrehungszahl.
Auf der Anzeige "Time/Prog" (A-3) wird die abgelaufene Zeit ab dem Datum angezeigt, an dem die Taste "Quick
Spin" gedrückt wurde (8).
Durch Freigabe der Taste "Quick Spin" (8) stoppt die Zentrifuge und die Laufzeit wird bis zum Öffnen des
Deckels angezeigt.
A-3
M10
A-1
M3
M4
1
11
12
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-13
Abbildung. 26
3.6.2 Die "STOP"-Taste
Mit der Taste "Stop" (10) (siehe Abbildung 27) können Sie den Durchlauf jederzeit unterbrechen. Nachdem Sie die
Taste gedrückt haben, wird sich die Zentrifuge mit der vorab ausgewählten Intensität bis zum Stillstand
verlangsamen.
Abbildung. 27
3.7 Feststellung eines Ungleichgewichts
Für den Fall, dass der Rotor nicht gleichmäßig beladen ist, wird der Antrieb sich bei der Beschleunigung
ausschalten. Der Rotor verlangsamt sich bis zum Stillstand.
Wenn in der Anzeige "Time/Prog" (A-3) das Wort "error” (M11) zusammen mit der Zahl "01” erscheint, ist der
Gewichtsunterschied zwischen den Proben zu groß. Verteilen Sie das Gewicht gleichmäßig.
Beladen Sie den Rotor wie in Kapitel 3.1.2 und 3.1.3 beschrieben.
Wenn auf der Anzeige "Time/Prog" (A-3) das Wort "error” zusammen mit der Zahl "02" (siehe Abbildung 28)
erscheint, könnte dies folgenden Grund haben: Der Schalter für Ungleichgewicht ist defekt.
Abbildung 28
10
A-3
M11
1
8
9
A-3
DE-14 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4. EINSTELLUNG
4.1 Grundanpassungen
4.1.1 Zugriff auf den Modus "Betriebsdaten"
Wenn Sie die Zentrifuge benutzen, können die folgenden Parameter eingestellt werden:
- Temperaturanzeige °C oder °F
- Akustisches Signal ein-/ausschalten
- Tastaturgeräusch ein-/ ausschalten
- Lautstärke-Vorauswahl des Klangsignals
- Melodieauswahl des Klangsignals "Ende des Durchlaufs"
Waehrend die Zentrifuge ausgeschaltet ist, drücken Sie gleichzeitig die Tasten "Time"(6) und "Door Open"
(7) und schalten Sie die Zentrifuge am Hauptschalter ein. Lassen Sie nun beide Tasten los und als ein Ergebnis
hiervon wird ein Anzeigetest von ungefähr 5 Sekunden durchgeführt. Alle Indikatoren werden gleichzeitig
erscheinen (siehe Abbildung 29).
Abbildung. 29
ACHTUNG:
Bitte beachten Sie, dass Sie das Programm wie unter Punkt 4.1.1 beschrieben eingeben ssen, um
die Anpassungen der Punkte 4.1.2 - 4.1.7 zu ändern. Nachdem Sie die Einstellungen gespeichert
haben, können Sie den normalen Programmiermodus wieder ändern, indem Sie die Zentrifuge
kurzzeitig ausschalten.
Alle geänderten Einstellungen müssen mit der Taste "Start"(9) bestätigt werden. Das Wort "Store"(12) erscheint
auf der Anzeige "RPM | RCF"(A-1) - Nur dann sind die Vorauswahlen gültig!
Abbildung. 30
6
7
A-1
12
9
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-15
4.1.2 Temperaturangabe
Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann
die Taste "Accel/Decel" (5). Auf der Anzeige "Acc/Dec" (A-2) erscheint das Wort "Service". Waehlen Sie nun
den Buchstaben "C" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen auf der Anzeige "RPM | RCF"
(A-1) die Worte "CELSI/temp" . Wenn Sie die Taste "RPM | RCF" (4) drücken, blinkt das Wort "CELSI" und Sie
können mit dem Anpassungsknopf (1) die Anzeige in Fahrenheit "FAREN" ändern (siehe Abbildung 31).
Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), ändern Sie zurück zum normalen
Programmiermodus, indem Sie die Zentrifuge kurzzeitig ausschalten.
Abbildung. 31
4.1.3 Signal ein-/ausschalten
Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann
die Taste "Accel/Decel"(5). Auf der Anzeige blinkt "Acc/Dec" (A-2) das Wort "Service" . Wählen Sie nun den
Buchstaben "L" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen die Worte "On Sound" auf der
Anzeige "RPM | RCF" (4). Wenn Sie die Taste "RPM | RCF" (4) nun drücken, blinkt das Wort "On" und Sie
können den Ton mit dem Anpassungsknopf (1) umschalten (siehe Abbildung 32).
Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), ändern Sie zurück zum normalen
Programmiermodus, indem Sie die Zentrifuge kurzzeitig ausschalten.
Abbildung. 32
4
5
1
4
5
1
A-2
A-1
DE-16 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4.1.4 Lautstärke-Vorauswahl des Tonsignals
Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann
die Taste "Accel/Decel" (5). Auf der Anzeige "Acc/Dec" (A-2) blinkt das Wort "Service" . Wählen Sie nun den
Buchstaben "U" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen auf der Anzeige "RPM | RCF" (A-1)
die Worte "Vol=0- 9/Sound" . Nach dem Drücken der Taste "RPM | RCF" (4), können Sie die gewünschte
Lautstaerke zwischen 0 (leise) und 9 (laut) mit dem Anpassungsknopf (1) anpassen (siehe Abbildung 33).
Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), können Sie zurückwechseln zum normalen
Programmiermodus, indem Sie die Zentrifuge für einen kurzen Zeitraum ausschalten.
Abbildung. 3
4.1.5 Liedauswahl für das Tonsignal am Ende des Durchlaufs
Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann
die Taste "Accel/Decel" (5). Auf der Anzeige blinkt "Acc/Dec" (A-2) das Wor t"Service" . Wählen Sie nun den
Buchstaben "G" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen auf der Anzeige "RPM | RCF" (A-1)
die Worte "SonGo/Sound" . Nach dem Drücken der Taste "RPM | RCF" (4) können Sie mit Hilfe des
Anpassungsknopfes (1) ein Lied auswählen. (Siehe Abbildung 34).
Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), können Sie zurückwechseln zum normalen
Programmiermodus, indem Sie die Zentrifuge kurzzeitig ausschalten.
Abbildung. 34
4.1.6 Tastaturgeräusch ein-/ ausschalten
Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann
die Taste "Accel/Decel" (5). Auf der Anzeige "Acc/Dec" (A-2) blinkt das Wort "Service" . Wählen Sie nun den
Buchstaben "B" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen auf der Anzeige "RPM | RCF" (A-1)
die Worte "ON/BEEP" . Nach dem Drücken der Taste "RPM | RCF" (4) können Sie das Tastaturgeräusch mit dem
Anpassungsknopf (1) ein- (On) oder ausschalten (Off). (Siehe Abbildung 39).
Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), können Sie zurückwechseln zum normalen
Programmiermodus, indem Sie die Zentrifuge kurzzeitig ausschalten.
A-1
A-2
4
5
1
4
5
1
A-2
A-1
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-17
Abbildung. 35
4.1.7 Betriebsdaten aufrufen
ACHTUNG:
Dies sollte nur durch einen fortgeschrittenen Benutzer oder Servicetechniker ausgeführt werden.
Im Modus "Grundanpassungen" können Sie die Betriebsdaten der Zentrifuge aufrufen. Fahren Sie
bitte wie unter Punkt 4.1.2 beschrieben fort, um in diesen Programmmodus einzutreten. Drücken Sie die Taste
"Accel/Decel" (5). Auf der Anzeige "Acc/Dec" (A-2) blinkt das Wort "Service" .
Mit dem Anpassungsknopf (1) können die verschiedenen Informationen erreicht werden:
A= vorherige Starts der Zentrifuge
H= vorherige Betriebsstunden
S= Softwareversion
r= Konvertersoftware
E= Liste der vorherigen Fehlermitteilungen
h= Laufzeit des Motors
Die Liste der letzten 99 Fehlermitteilugen kann aufgerufen werden, indem die Taste "RPM | RCF" (4) gedrückt wird,
und Sie können mit Hilfe des Anpassungsknopfes (1) durchscrollen. Die entsprechenden Fehlercodes erscheinen
in der Anzeige "RPM | RCF" (A-1). Siehe bitte "Tabelle 6: Fehlermitteilungen" (siehe ANHANG).
Um auf den normalen Programmmodus zu wechseln, schalten Sie die Zentrifuge für eine kurze Zeit aus.
Abbildung 36
4
1
5
A-2
A-1
4
5
1
A-2
A-1
DE-18 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
5. WARTUNG
5.1 Wartung und Reinigung
5.1.1 Allgemein
Pflege:
Die Wartung der Zentrifuge ist darauf beschränkt, den Rotor, die Rotorkammer und die Rotorzubehörteile sauber
zu halten sowie die Rotoreinsatzschrauben eines Ausschwenkrotors (falls vorhanden) regelmäßig zu schmieren.
Das geeignetste Schmiermittel ist das High TEF-Öl.
Schmiermittel, die Molycote und Graphit enthalten, sind nicht zugelassen.
Achten Sie vor allem auf die eloxierten Aluminiumteile. Sogar der leichteste Schaden kann zu einem Bruch der
Rotoren führen.
Für den Fall dass Rotoren, Behälter oder Rohrgestelle mit korrosiven Substanzen in Kontakt stehen, müssen die
betroffenen Stellen sorgfältig gereinigt werden.
Korrosive Substanzen sind zum Beispiel: Alkalien, Basen, Seifenlösungen, Alkaliamine, konzentrierte Säuren,
Lösungen, die Schwermetalle enthalten, wasserfreie, chlorhaltige Lösungsmittel, Salzlösungen, z.B. Salzwasser,
Phenol, halogenierte Kohlenwasserstoffe.
Reinigung - Einheiten, Rotoren, Zubehörteile
- Schalten Sie das Gerät aus und trennen Sie es von der Stromversorgung, bevor Sie mit irgendwelchen
Reinigungs- und Desinfektionsarbeiten beginnen. Schütten Sie keine Flüssigkeiten in das Innere des
Gehäuses.
- Sprühen Sie kein Desinfektionsmittel auf das Gerät.
- Eine gründliche Reinigung erfüllt nicht nur seinen Zweck hinsichtlich der Hygiene, sondern vermeidet auch
Korrosion aufgrund von Verschmutzung.
- Um eine Beschädigung von eloxierten Teilen wie z.B. Rotoren, Reduktionsplatten, usw. zu vermeiden,
benutzen Sie nur pH-neutrale Reinigungsmittel mit einem pH-Wert von 6-8 zur Reinigung. Basische
Reinigungsmittel (pH-Wert > 8) dürfen nicht benutzt werden.
- Stellen Sie nach der Reinigung sicher, dass alle Teile gründlich getrocknet sind, entweder per Hand oder in
einer Heißluftkammer (max. Temperatur + 50°C).
- Es ist erforderlich, die eloxierten Aluminiumteile regelmäßig mit einem Korrosionsschutz-Öl zu beziehen, um
ihre Lebensdauer zu erhöhen und die Korrosionsanfälligkeit zu reduzieren.
- Aufgrund von Feuchtigkeit oder nicht hermetisch verriegelten Proben können sich Kondensate bilden. Das
Kondensat muss von der Rotorkammer regelmäßig mit einem weichen Tuch entfernt werden.
Das Wartungsverfahren muss alle 10 bis 15 Durchläufe oder mindestens einmal pro Woche
wiederholt werden.
- Schließen Sie die Einheit an die Stromversorgung an, nachdem die Ausrüstung vollkommen trocken ist.
- Führen Sie keine Desinfektion mit UV-, Beta- und Gammastrahlen oder anderer hochenergetischer Strahlung
durch.
- Metallrotoren können autoklaviert werden.
- Rotordeckel und Adapter können ebenfalls autoklaviert werden (max. 121°C, 20 min).
- Die Reagenzglasgestelle sind aus PP und können nicht bei 134°C autoklaviert werden.
5.1.2 Reinigung und Desinfektion der Einheit
1. Öffnen Sie den Deckel bevor Sie die Einheit ausschalten. Trennen Sie sie von der Stromversorgung.
2. Öffnen Sie die Rotormutter, indem Sie den Rotorschlüssel im Uhrzeigersinn drehen.
3. Entfernen Sie den Rotor.
4. Benutzen Sie zur Reinigung und Desinfektion der Einheit und der Rotorkammer das oben genannte Reinigungsmittel.
5. Reinigen Sie alle zugänglichen Bereiche des Geräts und seine Zubehörteile, einschließlich des Stromkabel mit
einem feuchten Tuch.
6. Waschen Sie die Gummidichtungen und die Rotorkammer gründlich mit Wasser.
7. Schrubben Sie die Gummidichtungen mit Glyzerol oder Talkum trocken, um zu vermeiden, dass sie spröde
werden. Andere Komponenten der Einheit, z.B. der Deckelverschluss, die Motorwelle und der Rotor dürfen
nicht eingefettet werden.
8. Trocknen Sie die Motorwelle mit einem weichen, trockenen und fuselfreien Tuch.
9. Kontrollieren Sie die Einheit und die Zubehörteile auf Beschädigungen.
Stellen Sie sicher, dass die Zentrifuge abgeschaltet und trennen Sie die Einheit von der Stromversorgung.
Entfernen Sie dann den anhaftenden Staub von den Ventilationsschlitzen in der Zentrifuge, indem Sie eine weiche
Bürste verwenden. Wiederholen Sie dies mindestens alle 6 Monate.
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-19
5.1.3 Reinigung und Desinfektion des Rotors
1. Reinigen und desinfizieren Sie die Rotoren, Rotordeckel und Adapter mit dem oben genannten Reinigungsmittel.
2. Benutzen Sie eine Flaschenbürste, um die Rotorbohrungen zu reinigen und zu desinfizieren.
3. Spülen Sie die Rotoren, den Rotordeckel und den Adapter mit klarem Wasser ab. Vor allem die Bohrungen der
Winkelrotoren.
4. Um die Rotoren und Zubehörteile zu trocknen, legen Sie sie auf ein Handtuch. Stellen Sie die Winkelrotoren
mit den Bohrungen nach unten auf das Handtuch.
5. Trocknen Sie den Kegel mit einem weichen, trockenen und fuselfreien Tuch und halten Sie Ausschau nach
Beschädigungen. Fetten Sie den Rotorkegel nicht ein.
5.1.4 Desinfektion von Aluminiumrotoren
Im Falle von infektiösen Materialien, die in die Zentrifuge auslaufen, müssen der Rotor und die Rotorkammer direkt
nach dem Durchlauf desinfiziert werden. Rotoren können bei einer Maximaltemperatur von 121°C autoklaviert werden.
5.1.5 Desinfektion der PP-Rotoren
Autoklavieren
Die empfohlene Zeit fürs Autoklavieren: 1520 min bei 121°C (1 bar)
ACHTUNG:
Die Sterilisationszeit von 20 min. darf nicht überschritten werden. Eine wiederholte Sterilisation
wird zu einer Verminderung der mechanischen Widerstands des Kunststoffmaterials führen
Vor dem Autoklavieren müssen der PP-Rotor und der Adapter gründlich gereinigt werden, um ein Verbrennen von
Schmutzresten zu vermeiden.
Sie können die Auswirkungen einiger chemikalischen Reste auf Kunststoffmaterialien bei Umgebungstemperaturen
ignorieren. Aber bei den hohen Temperaturen beim Autoklavieren können diese Reste korrodieren und den
Kunststoff zerstören. Die Objekte müssen nach der Reinigung aber vor dem Autoklavieren gründlich mit
destilliertem Wasser abgespült werden. Reste von Reinigungsflüssigkeiten können Spalten, Aufhellungen und
Flecken hervorrufen.
Gassterilisation
Adapter, Flaschen und Rotoren kann mit Ethylenoxid gassterilisiert werden. Stellen Sie sichern, die Objekte nach
der Sterilisation und vor der erneuten Verwendung lüften.
ACHTUNG:
Da die Temperatur bei der Sterilisation ansteigen kann, dürfen Rotoren, Adapter und Flaschen
nicht geschlossen werden, und müssen vollkommen aufgeschraubt sein.
Chemische Sterilisation
Flaschen, Adapter und Rotoren können mit den üblichen flüssigen Desinfektionsmitteln benutzt werden.
ACHTUNG:
Bevor Sie eine andere Reinigungs- oder Dekontaminationsmethode als vom Hersteller
empfohlen anwenden, nehmen Sie mit dem Hersteller Kontakt auf, um sicherzustellen, dass dies
die Einheit oder den Rotor nicht beschädigen wird.
5.1.6 Glasbruch
Bei hohen G-Werten erhöht sich die Bruchrate von Reagenzgläsern aus Glas. Glassplitter müssen unverzüglich
vom Rotor, den Behälter und Adaptern sowie der Rotorkammer selbst entfernt werden. Feine Glassplitter werden
die Schutzbeschichtung eines Rotors kratzen und daher beschädigen. Wenn Glassplitter auf der Rotorkammer
verbleiben, wird sich aufgrund der Luftzirkulation Feinmetallstaub entstehen. Dieser sehr feine, schwarze
Metallstaub wird die Rotorkammer, den Rotor, die Behälter und die Proben erheblich verschmutzen.
Tauschen Sie, wenn erforderlich, Adapter, Rohre und Zubehörteile aus, um weitere Schäden zu vermeiden. Prüfen
Sie die Rotorlöcher regelmäßig auf Ablagerungen und Beschädigungen.
ACHTUNG:
Bitte prüfen sie die entsprechenden Spezifikation der Reagenzglaszentrifugen mit dem
Hersteller.
5.2 Lebenszeit der Rotoren, Behälter, Zubehörteile
Rotoren und Rotordeckel aus Aluminium oder rostfreiem Stahl verfügen über eine Betriebszeit von max. 7 Jahren
ab dem ersten Gebrauch.
Transparente Rotordeckel und Kappen aus PC oder PP sowie Rotoren, Reagenzglasgestelle und Adapter aus PP
verfügen über eine maximale Betriebsdauer von bis zu 3 Jahren ab dem ersten Gebrauch.
Bedingungen für die Betriebszeit: Ordnungsgemäße Anwendung, schadenfreier Zustand, empfohlene Pflege
DE-20 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
6. PROBLEMBEHEBUNG
6.1 Fehlermitteilung: Ursache / Lösung
Die Fehlermitteilungen werden aufgelistet, um Ihnen dabei zu helfen, mögliche Fehler schneller zu finden.
Die Diagnose, auf die in diesem Kapitel Bezug genommen wird, kann nicht immer zutreffen, da es sich nur um
theoretisch auftretende Fehler und Lösungen handelt.
Bitte halten Sie uns über alle möglichen auftretenden Fehler auf dem Laufenden, der in diesem Kapitel nicht
aufgeführt wird. Nur durch Ihre Informationen sind wir in der Lage, dieses Betriebshandbuch zu verbessern.
Vielen Dank im Voraus für Ihre Unterstützung.
6.2 Untersuchung der möglichen Fehlermitteilungen und ihrer Lösungen
6.2.1 Deckelfreigabe bei Stromausfall (Notfalldeckelfreigabe)
Im Falle eines Stromausfalls oder einer Fehlfunktion kann der Deckel der Zentrifuge manuell geöffnet werden, um
Ihre Proben herauszunehmen.
Für FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (motorgetriebenes Schloss)
Gehen Sie bitte wie folgt vor:
1. Schalten Sie die Zentrifuge aus und ziehen Sie das Stromkabel heraus, warten Sie, bis der Rotor still steht
(dies kann einige Minuten dauern)
2. Auf der linken Seite des Zentrifugengehäuses, gibt es einen Plastikstopper. Entfernen Sie diese Stopper,
dahinter befindet sich eine sechseckige Mutter.
3. Nehmen Sie den Steckschlüssel, stecken Sie ihn in das Loch und schließen Sie den Steckschlüssel mit der
sechseckigen Mutter (Siehe Abbildung 37).
4. Drehen Sie nun den Steckschlüssel bis zum Limit auf die rechte Seite (im Uhrzeigersinn).
ACHTUNG:
a) Drehen Sie nur bis zum Limit, ziehen Sie die Mutter nicht fest.
b) Öffnen Sie nun den Deckel der Zentrifuge.
c) Schalten Sie die Zentrifuge wieder ein, um die Arbeit wiederaufzunehmen.
Abbildung. 37
Für 5714
Bitte gehen Sie wie folgt vor (siehe Abbildung 38):
ACHTUNG:
Schalten Sie die Zentrifuge aus und ziehen Sie das Stromkabel heraus, warten Sie, bis der
Rotor still steht (dies kann einige Minuten dauern)
Auf der rechten Seite der Zentrifuge gibt es einen Plastikstopper (Abbildung 38). Entfernen sie
diesen Stopper, der am Deckelverschluss angeschlossen ist, horizontal vom Gehäuse, bis
sich der Zentrifugendeckel öffnet.
Öffnen Sie nun den Deckel der Zentrifuge
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-21
Abbildung. 38
6.2.2 Beschreibung des Fehlermitteilungssystems
Die Fehlermitteilung "Error" (M11) wird auf der "Time/Prog" (A-3)-Anzeige angezeigt (siehe Abbildung 39).
Detaillierte Informationen über mögliche Fehlermitteilungen sind enthalten in "Tabelle 6: Fehlermitteilungen"
(Siehe Anhang).
Abbildung. 39
7. ENTGEGENNAHME VON ZU REPARIERENDEN ZENTRIFUGEN
Gesundheitsrisiko von kontaminierter Ausrüstung, Rotoren und Zubehörteilen.
Für den Fall, dass die Zentrifuge zur Reparatur an den Hersteller zurückgesandt wird, beachten Sie bitte
Folgendes:
Die Zentrifuge muss vor der Versendung dekontaminiert und gereinigt werden, um Personen, Umwelt und Material
zu schützen.
Dekontaminationszertifikat für die Rücksendung von Gütern (Siehe ANHANG)
Wir behalten uns das Recht vor, kontaminierte Zentrifugen nicht anzunehmen.
Zudem gehen alle Kosten, die bei der Reinigung und Desinfektion der Einheiten entstehen zu Lasten des Kunden.
A-3
M11
DE-22 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
8. TRANSPORT, LAGERUNG UND ENTSORGUNG
8.1 Transport
Nehmen Sie den Rotor vor dem Transport heraus.
Transportieren Sie die Einheit nur in der Originalverpackung.
Installieren Sie das Transportschutzmaterial, um die Motorwelle zu schützen,
wenn Sie das Gerät über längere Strecken transportieren.
Air temperature rel. humidity Air pressure
General transportation -25 to 60 °C 10 to 75 % 30 to 106 kPa
8.2 Lagerung
Während der Lagerung der Zentrifuge müssen die folgenden
Umweltbedingungen eingehalten werden:
Air temperature rel. Humidity Air pressure
in transport packaging -25 to 55 °C 10 to 75 % 70 to 106 kPa
8.3 Transport, Installation, Übertragung und Entsorgung der Zentrifuge
FC5714/FC5718/FC5718R/FC5816/FC5816R
Diese Anweisungen ergänzen die vorherigen allgemeinen
Anweisungen in Kapitel 8 und ersetzen sie nicht
8.3.1 Transport
Bitte transportieren Sie das Gerät in der Originalverpackung.
Die Zentrifuge sollte immer unter Verwendung von mechanischem
Transportgerät transportiert werden.
8.3.2 Installation
Öffnen des Kartons und Heben des Geräts
1. Schneiden Sie das Klebeband ab.
2. beschrieben.Öffnen Sie alle 4 Klappen des Kartons
3. Entfernen Sie die Zubehörteile
4. Heben Sie die Zentrifuge vorsichtig aus dem Karton.
WARNUNG: Hebegefahr. Das Heben durch eine einzige Person kann Unfälle verursachen.
Benutzen Sie ein mechanisches Hebegerät oder verknüpfen Sie Hebeverfahren, wenn Sie die
Ausrüstung heben oder bewegen.
Platzieren Sie das Gerät auf einem stabilen, horizontalen und nicht
vibrierenden Labortisch.
1. Entfernen Sie das vordere und hintere Schutzmaterial
2. Entfernen Sie die Plastikfolie.
3. Beachten Sie den Mindestabstand von 30 cm zu benachbarten Geräten an
den Seiten und von der Rückseite bis zur Wand.
4. Installieren Sie das Gerät an einem gut gelüfteten Standort, der vor
direktem Sonnenlicht geschützt ist, um eine Überhitzung zu vermeiden.
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R /FC5916/FC5916R DE-23
Schließen Sie das Gerät an
1. Warten Sie nach der Installation ungefähr vier Stunden, bevor Sie die
Zentrifuge einschalten, um so eine Beschädigung des Kompressors zu
vermeiden.
2. Prüfen Sie, dass die Netzspannung und die Frequenz den
Anforderungen auf der Namensplatte des Geräts (siehe Rückseite des
Geräts) übereinstimmen, und schließen Sie das Gerät dann an die
Stromversorgung an.
Entfernen Sie das Transportschutzmaterial von der Rotorkammer.
1. Schalten Sie das Gerät am Hauptstromschalter ein.
2. Öffnen Sie den Zentrifugendeckel, indem Sie die Öffnen-Taste drücken
3. Entfernen Sie das Transportschutzmaterial
4. Platzieren Sie den Rotor vertikal auf dem Motorschaft.
5. Drehen Sie die Rotormutter, indem Sie den Rotorschlüssel im Uhrzeigersinn drehen, bis die Rotormutter
festgezogen ist.
Das Gerät ist nun gebrauchsfertig
Bewahren Sie die Verpackung und jegliches Transportmaterial für die
Versendung des Geräts zu einem späteren Zeitpunkt auf.
8.3.3 Verpackung
Packen Sie die Zentrifuge in umgekehrter Reihenfolge.
8.3.4 Weitergabe des Geräts
Wenn Sie die Ausrüstung an Dritte weitergeben, stellen Sie bitte sicher, dass Sie auch diese Betriebsanleitung
beifügen.
DE-24 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9. TECHNISCHE DATEN
9.1 Spezifikationen
9.1.1 Zentrifuge FC5714
Modell
FC5714
Geschwindigkeitsreichweite
200 rpm -14000 rpm;10 rpm/Satz
Maximales RCF
18624 x g;10 x g/Satz
Maximale Kapazität(Rotor)
4x100ml
Temperaturreichweite (Nicht verfügbar)
Luftkühlung
Laufzeit
10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft
Lärmpegel (je nach Rotor)
63 ± 2 dB(A)
Zulässige Dichte bei
Maximalgeschwindigkeit
1,2 g/ml
Zulässige kinetische Energie
5595 Nm
Netzanschluss AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Spannungsschwankung
± 10 %
± 10 %
Aktueller Verbrauch
1,3 A
2,4 A
Stromverbrauch
240 W
300 W
Ausmaße (B × T × H)
362 x 493 x 330 mm
14,3 x 19,4 x 13,0 in
Nettogewicht (ohne Rotor)
30 kg
66 lb
Versandausmaße (B × T × H)
580 x 490 x 460 mm
22,8 x 19,3 x 18,1 in
Versandgewicht (ohne Rotor)
32,5 kg
72 lb
Umgebungsbedingungen (EN/IEC 61010-1)
Umwelt
nur zur Innenanwendung
Höhe
Bis zu einer Höhe von 2000m benutzbar
Umgebungstemperatur
2°C bis 35 °C
Max. relative Luftfeuchtigkeit
Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis
zu 31°C,
linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend
bis 35°C.
Überspannungskategorie (IEC 60364-4-
443)
II
Kontaminationsgrad
2
Schutzgrad
I
Nicht geeignet für die Anwendung in gefährlichen Umgebungen.
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-25
9.1.2 Zentrifuge FC5718
Modell
FC5718
Geschwindigkeitsreichweite
200 rpm -18000 rpm;10 rpm/Satz
Maximales RCF
23542 x g;10 x g/Satz
Maximale Kapazität (Rotor)
4x100ml
Temperaturreichweite (Nicht verfügbar)
Luftkühlung
Laufzeit Lärmpegel
10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft
Lärmpegel (je nach Rotor)
60 ± 2 dB(A)
Zulässige Dichte bei
Maximalgeschwindigkeit
1,2 g/ml
Zulässige kinetische Energie
16672 Nm
Netzanschluss AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Spannungsschwankung
± 10 %
± 10 %
Aktueller Verbrauch
2.0 A
4.0 A
Stromverbrauch
455 W
475 W
Ausmaße (B × T × H)
408 x 499 x 351 mm
16,1 x 19,7 x 13,8 in
Nettogewicht (ohne Rotor)
43 kg
95 lb
Versandausmaße (B × T × H)
650 x 520 x 490 mm
25,6 x 20,5 x 19,3 in
Versandgewicht (ohne Rotor)
53 kg
117 lb
Umgebungsbedingungen (EN/IEC 61010-1)
Umwelt
nur zur Innenanwendung
Höhe
Bis zu einer Höhe von 2000m benutzbar
Umgebungstemperatur
2°C bis 35 °C
Max. relative Luftfeuchtigkeit
Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis zu
31°C,
linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend bis
35°C.
Überspannungskategorie (IEC 60364-4-
443)
II
Kontaminationsgrad
2
Schutzgrad
I
Nicht geeignet für die Anwendung in gefährlichen Umgebungen.
DE-26 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R /FC5916/FC5916R
9.1.3 Zentrifuge FC5718R
Modell
FC5718R
Geschwindigkeitsreichweite
200 rpm -18000 rpm;10 rpm/Satz
Maximales RCF
23542 x g;10 x g/Satz
Maximale Kapazität(Rotor)
4x100ml
Temperaturreichweite(Digital)
-20~40
Laufzeit
10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft
Lärmpegel (je nach Rotor)
60 ± 2 dB(A)
Zulässige Dichte bei
Maximalgeschwindigkeit
1,2 g/ml
Zulässige kinetische Energie
25111 Nm
Netzanschluss AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Spannungsschwankung
± 10 %
± 10 %
Aktueller Verbrauch
3,0 A
6,0 A
Stromverbrauch
660 W
660 W
Ausmaße (B × T × H)
407 x 731 x 359 mm
16,0 x 28,8 x 14,1 in
Nettogewicht (ohne Rotor)
60 kg
132 lb
Versandausmaße (B × T × H)
840 x 640 x 590 mm
33,1 x 25,2 x 23,2 in
Versandgewicht (ohne Rotor)
77 kg
170 lb
Umgebungsbedingungen (EN/IEC 61010-1)
Umwelt
nur zur Innenanwendung
Höhe
Bis zu einer Höhe von 2000m benutzbar
Umgebungstemperatur
2°C bis 35 °C
Max. relative Luftfeuchtigkeit
Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis zu
31°C,
linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend bis
35°C.
Überspannungskategorie (IEC 60364-4-
443)
II
Kontaminationsgrad
2
Schutzgrad
I
Nicht geeignet für die Anwendung in gefährlichen Umgebungen.
9.1.4 Centrifuge FC5816
Modell
FC5816
Geschwindigkeitsreichweite
200 rpm -6000 rpm;50 rpm/Satz
Maximales RCF
4427 x g;10 x g/Satz
Maximale Kapazität(Rotor)
6 x 250ml
Temperaturreichweite (Nicht verfügbar)
Luftkühlung
Laufzeit
10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft
Lärmpegel (je nach Rotor)
61 +2 dB(A)
Zulässige Dichte bei
Maximalgeschwindigkeit
1,2 g/ml
Zulässige kinetische Energie
34363 Nm
Netzanschluss AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Spannungsschwankung
± 10 %
± 10 %
Aktueller Verbrauch
2,4 A
4,2 A
Stromverbrauch
530 W
520 W
Ausmaße (B × T × H)
446 x 538 x 354 mm
17,6 x 21,2 x 13,9 in
Nettogewicht (ohne Rotor)
52 kg
115 lb
Versandausmaße (B × T × H)
840 x 640 x 590 mm
33,1 x 25,2 x 23,2 in
Versandgewicht (ohne Rotor)
77 kg
170 lb
Umgebungsbedingungen (EN/IEC 61010-1)
Umwelt
nur zur Innenanwendung
Höhe
Bis zu einer Höhe von 2000m benutzbar
Umgebungstemperatur
2°C bis 35 °C
Max. relative Luftfeuchtigkeit
Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis zu
31°C,
linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend bis
35°C.
Überspannungskategorie (IEC 60364-4-
443)
II
Kontaminationsgrad
2
Schutzgrad
I
Nicht geeignet für die Anwendung in gefährlichen Umgebungen.
DE-28 Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.5 Zentrifuge FC5816R
Modell
FC5816R
Geschwindigkeitsreichweite
200 rpm -16000 rpm;10 rpm/Satz
Maximales RCF
24325 x g;10 x g/Satz
Maximale Kapazität(Rotor)
6 x 250ml
Temperaturreichweite(Digital)
-20~40
Laufzeit
10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft
Lärmpegel (je nach Rotor)
63 ± 2 dB(A)
Zulässige Dichte bei
Maximalgeschwindigkeit
1,2 g/ml
Zulässige kinetische Energie
34363 Nm
Netzanschluss AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Spannungsschwankung
± 10 %
± 10 %
Aktueller Verbrauch
3,7 A
7,8 A
Stromverbrauch
785 W
850 W
Ausmaße (B × T × H)
723 x 538 x 354 mm
28,5 x 21,2 x 13,9 in
Nettogewicht (ohne Rotor)
77 kg
170 lb
Versandausmaße (B × T × H)
840 x 640 x 590 mm
33,1 x 25,2 x 23,2 in
Versandgewicht (ohne Rotor)
87 kg
192 lb
Umgebungsbedingungen (EN/IEC 61010-1)
Umwelt
nur zur Innenanwendung
Höhe
Bis zu einer Höhe von 2000m benutzbar
Umgebungstemperatur
2°C bis 35 °C
Max. relative Luftfeuchtigkeit
Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis zu
31°C,
linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend
bis 35°C.
Überspannungskategorie (IEC 60364-4-
443)
II
Kontaminationsgrad
2
Schutzgrad
I
Nicht geeignet für die Anwendung in gefährlichen Umgebungen.
9.1.6 Centrifuge FC5916
Modell
FC5916
Geschwindigkeitsbereich
200 U/min -16000 U/min;10 U/min/Satz
Maximale RCF
24325 x g;10 x g/Satz
Maximale Kapazit(Rotor)
4x750ml
Temperaturbereich(N/A)
Luftkühlung
Laufzeit
10 sec bis 99 Std. 59 min 59 sec oder kontinuierlich
Geuschpegel (abhängig vom Rotor)
63 ± 2 dB(A)
Zulässige Dichte bei maximaler
Geschwindigkeit
1.2 g/ml
Zulässige kinetische Energie
60629 Nm
Netzanschluss AC
230 V ~ 50/60
Hz
120 V ~ 50/60 Hz
Spannungsschwankung
± 10 %
± 10 %
Aktueller Stromverbrauch
2.8 A
5.6 A
Leistungsaufnahme
640 W
680 W
Maße (B × T × H)
540 x 670 x 390 mm
21.3 x 26.4 x 15.4 in
Nettogewicht (ohne Rotor)
85 kg
187 lb
Maße für den Versand (B × T × H)
780 x 670 x 590 mm
30.7 x 26.4 x 23.3 in
Transportgewicht (ohne Rotor)
98 kg
216 lb
Umgebungsbedingungen (EN/IEC 61010-1)
Bereich
Nur für den Innenbereich
Höhe
Einsatz bis zu einer he von 2000 m
Umgebungstemperatur
C bis zu 35°C
Max. relative Luftfeuchtigkeit
Max. relative Luftfeuchtigkeit 80 % für Temperaturen bis 31°C,
linear abnehmend auf 50 % relative Luftfeuchtigkeit bis 35°C.
Überspannungskategorie (IEC 60364-4-
443)
II
Grad der Verschmutzung
2
Schutzklasse
I
Nicht geeignet für den Einsatz in gefährlicher Bereiche.
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-29
DE-30
9.1.7 Centrifuge FC5916R
Modell FC5916R
Geschwindigkeitsbereich
200 U/min -16000 U/min;10 U/min/Satz
Maximale RCF
26331 x g;10 x g/Satz
Maximale Kapazit(Rotor)
4x750ml
Temperaturbereich(Digital)
-20~40
Laufzeit
10 sec bis 99 Std. 59 min 59 sec oder kontinuierlich
Geuschpegel (abhängig vom Rotor)
63 ± 2 dB(A)
Zulässige Dichte bei maximaler
Geschwindigkeit
1.2 g/ml
Zulässige kinetische Energie
54458 Nm
Netzanschluss AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Spannungsschwankung
± 10 %
± 10 %
Aktueller Stromverbrauch
7.2 A
20 A
Leistungsaufnahme
1630 W
1750 W
Maße (B × T × H)
730 x 670 x 390 mm
28.7 x 26.4 x 15.4 in
Nettogewicht (ohne Rotor)
118 kg
260 lb
Maße für den Versand (B × T × H)
900 x 750 x 560 mm
40.0 x 29.5 x 22.0 in
Transportgewicht (ohne Rotor)
137 kg
137 kg
Umgebungsbedingungen (EN/IEC 61010-1)
Bereich
Nur für den Innenbereich
Höhe
Einsatz bis zu einer he von 2000 m
Umgebungstemperatur
C bis zu 35°C
Max. relative Luftfeuchtigkeit
Max. relative Luftfeuchtigkeit 80 % für Temperaturen bis 3C,
linear abnehmend auf 50 % relative Luftfeuchtigkeit bis 35°C.
Überspannungskategorie (IEC 60364-4-
443)
II
Grad der Verschmutzung
2
Schutzklasse
I
Nicht geeignet für den Einsatz in gefährlicher Bereiche.
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.2 Zeichnungen und Ausmaße
D
H
W
W
H
D
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-31
DE-32
Modell
B (mm / in.)
T (mm / in.)
H (mm / in.)
FC5714
362 / 14.3
493 / 19.4
330 / 13.0
FC5718
408 / 16.1
499 / 19.7
351 / 13.8
FC5718R
408 / 16.1
731 / 28.8
359 / 14.1
FC5816
446 / 17.6
538 / 21.2
354 / 13.9
FC5816R
723 / 28.5
538 / 21.2
354 / 13.9
FC5916
540 / 21.3
670 / 26.4
390 / 15.4
FC5916R
730 / 28.7
670 / 26.4
390 / 15.4
10. COMPLIANCE
Die Erllung der folgenden Standards wird durch die entsprechende Kennzeichnung auf dem Produkt
angegeben.
Kennzeichn
ung
Standard
Dieses Produkt entspricht den geltenden, harmonisierten Standards der EU-
Direktiven 2011/65/EU (RoHS), 2014/30/EU (EMC) und 2014/35/EU (LVD).
The EU-Konformitätserklärung ist online verfüg bar unter
www.ohaus.com/ce.
Entsorgung
In Übereinstimmung mit der Euroischen Direktive 2012/19/EU über Waste
Electrical and Electronic Equipment (WEEE), darf dieses Gerät nicht mit dem
uslichen Abfall entsorgt werden. Dies gilt auch für Län der außerhalb der EU, in
Bezug auf ihre spezifischen Anforderungen.
Bitte entsorgen Sie dieses Produkt entsprechend lokalen Bestimmungen an
Sammelpunkten, die für elektrische und elektronische Ausrüstungen spezifiziert
sind.
Wenn Sie Fragen haben, nehmen Sie bitte mit der verantwortlichen Behörde oder
dem Händler Kontakt auf, von dem Sie dieses Gerät gekauft haben.
Sollte dieses Gerät an andere Parteien weitergegeben werden (für den privaten
oder beruflichen Gebrauch), muss auch auf den Inhalt dieser Bestimmung
verwiesen werden.
Für Anweisungen zur Entsorgung in Europa, siehe www.ohaus.com/weee.
Vielen Dank für Ihren Beitrag zum Umweltschutz.
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-33
11. ANHANG
TABELLE 1: ZULÄSSIGES NETTOGEWICHT
TABELLE 2: NIEDRIGSTE TEMPERATUREN UND MAX.-GESCHWINDIGKEIT
TABELLE 3: MAX. GESCHWINDIGKEIT UND RCF-WERTE R ZUSSIGE ROTOREN
TABELLE 4: BESCHLEUNIGUNGS- UND VERLANGSAMUNGSZEITEN
TABELLE 5: FEHLERMITTEILUNGEN
TABELLE 6 RADIUSKORREKTUR
TABELLE 7: CKNAHMEFORMULAR / DEKONTAMINATIONSZERTIFIKAT
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-34
11.1 Table 1: ZULÄSSIGES NETTOGEWICHT
Rotor-Nummer
Rotorbeschreibung
Zulä ssiges
Nettogewicht
Maximale Drehzahl ( x1000 U/min)
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
16.0
16.0
30314820
Rotor, Sw ing out, 4x250ml, ID
4x 536 g
4.5
4.5
30314821
Rotor, Angle, 6x250ml FB, ID
6x 355 g
8.0
8.0
8.0
8.0
30314822
Rotor, Sw ing out, 4x100ml, ID
4x 340 g
4.5
4.5
4.5
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
4x 465 g
4.0
5.0
5.0
30314824
Rotor, Sw ing out, 2x3MTP, ID
2x 310 g
4.5
4.5
4.5
4.5
4.5
4.5
4.5
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
6x 140 g
11.0
13.5
30314826
Rotor, Angle, 6x 85ml RB, ID
6x 140 g
9.0
9.0
11.0
13.0
11.0
11.0
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
4x 140 g
12.0
12.0
12.0
12.0
15.0
15.0
30314828
Rotor, Sw ing out, 16x 50ml, ID
4x 557 g
4.5
4.5
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-35
30314829
Rotor, Angle, 10x50ml FA, ID
10x 74 g
7.5
7.5
9.0
10.5
10.0
10.0
30314830
Rotor, Angle, 6x50ml RB/FA, ID
6x 72 g
6.0
6.0
6.0
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
6x 94 g
12.0
12.0
13.0
13.0
13.0
13.0
30314832
Rotor, Angle, 30x15ml RB/FA, ID
30x 32 g
4.5
4.5
4.5
4.5
4.5
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
20x 18 g
12.0
12.0
12.0
12.0
12.0
12.0
30314834
Rotor, Angle, 12x15ml RB/FA, ID
12x 25 g
6.0
6.0
6.0
30314835
Rotor, Angle, 44x 1.5/2.0ml, ID
44x 3.8 g
13.5
13.5
15.0
16.0
16.0
16.0
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
30x 3.4 g
12.0
13.0
14.0
15.0
15.0
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
15.0
15.0
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
12x 3.4 g
18.0
18.0
30314840
Rotor, Angle, 64x0.5ml, ID
64x 2.3 g
13.5
13.5
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
4x 3.5 g
15.0
15.0
15.0
15.0
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
4x 995 g
4.0
4.0
30553086
Rotor Angle 4x 500ml D69mm ID
4x 708 g
8.0
8.0
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-36
1.2 Table 2: NIEDRIGSTE TEMPERATUREN UND MAX.-GESCHWINDIGKEIT
Rotor-Nummer
Rotorenbeschreibung
Modell
Maximale
Geschwindigkeit
N-max
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314821
Rotor, Angle, 6x250ml FB, ID
FC5916R
8000 rpm
-1 ° C
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5718R
4500 rpm
-7 ° C
FC5816R
4500 rpm
-7 ° C
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5718R
5000 rpm
-3 ° C
FC5816R
5000 rpm
-3 ° C
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5718R
4500 rpm
-6 ° C
FC5816R
4500 rpm
-6 ° C
FC5916R
4500 rpm
-8 ° C
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718R
13500 rpm
17 ° C
FC5816R
13500 rpm
17 ° C
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718R
9000 rpm
-5 ° C
FC5816R
9000 rpm
-5 ° C
FC5916R
13000 rpm
8 ° C
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718R
12000 rpm
5 ° C
FC5816R
12000 rpm
5 ° C
FC5916R
16000 rpm
7 ° C
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718R
7500 rpm
9 ° C
FC5816R
7500 rpm
9 ° C
FC5916R
10500 rpm
2 ° C
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5718R
6000 rpm
-8 ° C
FC5816R
6000 rpm
-8 ° C
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718R
12000 rpm
-5 ° C
FC5816R
12000 rpm
-5 ° C
FC5916R
13000 rpm
-4 ° C
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5718R
4500 rpm
-9 ° C
FC5816R
4500 rpm
-9 ° C
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718R
12000 rpm
2 ° C
FC5816R
12000 rpm
2 ° C
FC5916R
12000 rpm
-2 ° C
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5718R
6000 rpm
-11.5 ° C
FC5816R
6000 rpm
-11.5 ° C
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718R
13500 rpm
1 ° C
FC5816R
13500 rpm
1 ° C
FC5916R
16000 rpm
3 ° C
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5718R
14000 rpm
9 ° C
FC5816R
14000 rpm
9 ° C
FC5916R
15000 rpm
3 ° C
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718R
18000 rpm
-2 ° C
FC5816R
18000 rpm
-2 ° C
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718R
13500 rpm
2 ° C
FC5816R
13500 rpm
2 ° C
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718R
15000 rpm
-1 ° C
FC5816R
15000 rpm
-1 ° C
FC5916R
15000 rpm
-2 ° C
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916R
4500 rpm
4 ° C
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916R
8000 rpm
0 ° C
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Alle Temperaturangaben beziehen sich auf eine Raumtemperatur von 23 ° C. Durch Überschreiten dieses Wertes oder direkte Sonneneinstrahlung auf die
Zentrifuge können diese Werte nicht eingehalten werden.
DE-37
1.3 Table 3: MAX. GESCHWINDIGKEIT UND RCF-WERTE FÜR ZULÄSSIGE ROTOREN
Rotor-Nummer
Rotorenbeschreibung
Modell
Maximale Geschwindigkeit
U/MIN
RCF
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
4500
3780 x g
FC5816R
4500
3780 x g
30314821
Rotor, Angle, 6x250ml FB, ID
FC5816
8000
10016 x g
FC5816R
8000
10016 x g
FC5916
8000
10016 x g
FC5916R
8000
10016 x g
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5714
4500
3350 x g
FC5718
4500
3350 x g
FC5718R
4500
3350 x g
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
4000
2486 x g
FC5718
5000
3885 x g
FC5718R
5000
3885 x g
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5714
4500
2716 x g
FC5718
4500
2716 x g
FC5718R
4500
2716 x g
FC5816
4500
2716 x g
FC5816R
4500
2716 x g
FC5916
4500
2716 x g
FC5916R
4500
2716 x g
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
11000
13932 x g
FC5718R
13500
20984 x g
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
9000
10413 x g
FC5718R
9000
10413 x g
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-38
FC5816
11000
15555 x g
FC5816R
13000
21726 x g
FC5916
11000
15555 x g
FC5916R
13000
21726 x g
30314827
Rotor, Angle, 4x 85ml RB, ID, Hi
FC5718
12000
14809 x g
FC5718R
12000
14809 x g
FC5816
12000
14809 x g
FC5816R
12000
14809 x g
FC5916
15000
23140 x g
FC5916R
16000
26328 x g
30314828
Rotor, Sw ing out, 16x50ml, ID
FC5816
4500
3735 x g
FC5816R
4500
3735 x g
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718
7500
8174 x g
FC5718R
7500
8174 x g
FC5816
9000
11771 x g
FC5816R
10500
16022 x g
FC5916
10000
14532 x g
FC5916R
10500
16022 x g
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
12000
13522 x g
FC5718R
12000
13522 x g
FC5816
13000
15869 x g
FC5816R
13000
15869 x g
FC5916
13000
15869 x g
FC5916R
13000
15869 x g
30314832
Rotor, Angle, 30x 15ml RB/FA, ID
FC5714
4500
2830 x g
FC5718
4500
2830 x g
FC5718R
4500
2830 x g
FC5816
4500
2830 x g
FC5816R
4500
2830 x g
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
12000
15775 x g
FC5718R
12000
15775 x g
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-39
FC5816
12000
15775 x g
FC5816R
12000
15775 x g
FC5916
12000
15775 x g
FC5916R
12000
15775 x g
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718
13500
17113 x g
FC5718R
13500
17113 x g
FC5816
15000
21128 x g
FC5816R
16000
24039 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5714
12000
15131 x g
FC5718
13000
17758 x g
FC5718R
14000
20595 x g
FC5916
15000
23643 x g
FC5916R
15000
23643 x g
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718
18000
23542 x g
FC5718R
18000
23542 x g
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
13500
16298 x g
FC5718R
13500
16298 x g
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
15000
15343 x g
FC5718R
15000
15343 x g
FC5916
15000
15343 x g
FC5916R
15000
15343 x g
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-40
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916
4000
3434 x g
FC5916R
4500
4346 x g
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916
8000
10367 x g
FC5916R
8000
10367 x g
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-41
11.4 Table 4: BESCHLEUNIGUNGS- UND VERLANGSAMUNGSZEITEN
Artikelnummer
Rotorbeschreibung
Modell
Beschleunigung* Zeit in
sek
Abbremsung* Zeit in
sek
level 0
level 9
level 0
level 9
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5714
200
25
230
35
FC5718
240
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
20
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
309
34
458
36
FC5816R
309
34
458
36
30314821
Rotor, Angle, 6x250ml FB, ID
FC5816
664
130
2906
92
FC5816R
664
130
2906
92
FC5916
573
66
1903
84
FC5916R
573
66
1903
84
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5714
110
14
170
19
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
110
14
170
17
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5714
230
25
340
26
FC5718
160
20
360
15
FC5718R
160
20
360
15
FC5816
452
43
616
38
FC5816R
452
43
616
38
FC5916
249
27
488
23
FC5916R
249
27
488
23
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
400
40
960
40
FC5718R
500
60
1260
50
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
360
40
1050
40
FC5718R
360
40
1050
40
FC5816
697
85
2313
70
FC5816R
2313
70
1630
76
FC5916
463
48
1654
46
FC5916R
549
69
1307
67
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718
300
60
820
40
FC5718R
300
60
820
40
FC5816
506
60
1745
49
FC5816R
506
60
1745
49
FC5916
448
50
1251
45
FC5916R
448
50
1251
45
30314828
Rotor, Sw ing out, 16x50ml, ID
FC5816
311
34
387
36
FC5816R
311
34
387
36
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718
380
63
796
37
FC5718R
380
62
1069
36
FC5816
753
115
2395
72
FC5816R
753
115
2395
72
FC5916
480
60
1747
68
FC5916R
480
60
1747
68
30314830
Rotor, Angle, 6x 50ml RB/FA, ID
FC5714
101
11
206
14
FC5718
89
13
239
11
FC5718R
90
12
463
11
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-42
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
446
48
1323
49
FC5816R
446
48
1323
49
FC5916
264
28
921
32
FC5916R
264
28
921
32
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5714
157
19
370
20
FC5718
160
15
380
10
FC5718R
160
15
380
10
FC5816
149
25
985
20
FC5816R
149
25
985
20
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
512
54
1439
47
FC5816R
512
54
1439
47
FC5916
305
32
988
37
FC5916R
305
32
988
37
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
102
11
167
14
FC5718
88
13
222
11
FC5718R
88
12
433
11
30314835
Rotor, Angle, 44x 1.5/2.0ml, ID
FC5718
196
22
204
18
FC5718R
196
22
204
18
FC5816
257
29
436
28
FC5816R
274
31
485
29
FC5916
233
25
283
30
FC5916R
233
25
500
25
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5714
250
27
280
34
FC5718
200
25
360
25
FC5718R
210
25
360
30
FC5916
221
23
561
30
FC5916R
221
23
561
30
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5714
200
22
230
35
FC5718
210
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
30
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718
240
25
210
20
FC5718R
240
25
210
20
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
150
17
170
12
FC5718R
150
17
170
12
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
130
16
130
12
FC5718R
130
16
130
12
FC5916
100
12
201
12
FC5916R
100
12
201
12
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916
483
47
1287
49
FC5916R
551
55
1501
54
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916
575
73
2317
82
FC5916R
575
73
2317
82
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
* Hinweis: Beschleunigt von 0 auf Vmax; bremst von Vmax auf 0 ab.
DE-43
11.5 Tabelle 5: FEHLERMITTEILUNGEN
Error-No.:
Descri pti on
1
Imbalance arose
2
Imbalance sensor is defectiv e
4
Imbalance switch has been activated for l onger than 5 secon ds
8
T ransponder i n the rotor i s defectiv e
11
T emperatu r e sensor is defective
12
Chamber over temperature
14
Leap of speed i s too big between two mesaurem e nts
CLOSE lid
33
Open lid while motor is running
34
Lid contact defectiv e
38
Lid motor is bl ocked
40
Communicatio n wi th frequen cy converter distrubed duri ng start
41
Communicatio n wi th frequen cy converter distrubed duri ng stop
42
Short circuit in the frequency converter
43
Undervoltag e frequen cy converter
44
Overvoltag e frequen cy converter
45
Over temperatu re frequency converter
46
Over temperatu re motor
47
Over current frequen cy conver ter
48
T i meout between control unit and frequency converter
49
Other error frequency converter
55
Overspeed
70
T i meout between controler and RS232 i nterface
99
Rotor i s not allowed i n this centri fuge
FALSE
Inserted rotor does not exi st i n the programm
rotor no
Rotor i s not detected
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-44
11.6 Table 6: RADIUSKORREKTUR
Rotor
Rotorbeschreibun
g
Buchse/Ad
apter
Buchse/Adapter Beschreibung
Radius
(cm)
Korrektur
(cm)
Kompatibel
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle,
24x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314820
Rotor, Sw ing out,
4x 250ml, ID
Blank
NO Bucket/Adapter
16.7
30314903
Rack, 1x250ml D62mm FB, 2/pk
15.7
-1.0
30314904
Rack, 1x100ml D41mm RB, 2/pk
15.9
-0.8
30314905
Rack, 3x50ml D29mm RB, 2/pk
16.3
-0.4
30314906
Rack, 3x50ml D29mm FA, 2/pk
16.7
0.0
30314907
Rack, 1x50ml D34mm RB, 2/pk
15.8
-0.9
30314908
Rack, 2x50ml D29mm Rim, 2/pk
15.8
-0.9
30314909
Rack, 1x50ml D29mm Rim, 2/pk
16.1
-0.6
30314910
Rack, 5x25ml D24.5mm RB, 2/pk
16.1
-0.6
30314911
Rack, 3x25ml D24.5mm Rim, 2/pk
15.5
-1.2
30314912
Rack, 9x15ml D17mm RB, 2/pk
16.3
-0.4
30314913
Rack, 7x15ml D17mm FA, 2/pk
16.4
-0.3
30314914
Rack, 7x15ml D17mm Rim, 2/pk
16.1
-0.6
30314915
Rack, 8x10ml D16mm RB, 2/pk
16.3
-0.4
30314916
Rack, 14x5-7ml D13mm RB, 2/pk
15.9
-0.8
30314917
Rack, 10x1.6-7ml D13mm Vac, 2/pk
15.9
-0.8
30304367
Rack, 8x4-10ml D16mm Vac, 2/pk
16.3
-0.4
30304368
Adapter, 9x 1.5/2.0ml D11mm, 2/pk
15.7
-1.0
30314821
Rotor, Angle,
6x 250ml FB, ID
Blank
NO Bucket/Adapter
14.1
30304369
Adapter, 8x 1.5ml D11mm, 2/pk
13.2
-0.9
30304370
Adapter, 5x 10ml D16mm RB, 2/pk
13.3
-0.8
30304371
Adapter, 4x 15ml D17mm FA, 2/pk
13.0
-1.1
30304372
Adapter, 2x 30ml D26mm RB, 2/pk
12.5
-1.6
30304373
Adapter, 1x 50ml D29mm RB, 2/pk
12.0
-2.1
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-45
30304374
Adapter, 1x 50ml D30mm FA, 2/pk
11.7
-2.4
30559412
Adapter 15x 1.5/2.0ml D11mm MTBx2
V1
30559413
Adapter 1x 250ml D60mm CBCx3 V1
11.8
-0.5
30559414
Adapter 1x 175/225ml D61mm FAx 2
V1
12.8
-2.3
30314822
Rotor, Sw ing out,
4x 100ml, ID
Blank
NO Bucket/Adapter
14.8
30314842
Rack, 1x100ml D46mm RB, 2/pk
14.1
-0.7
30314843
Rack, 1x85ml D38mm RB, 2/pk
14.1
-0.7
30314844
Rack, 1x50ml D34mm RB, 2/pk
14.2
-0.6
30314845
Rack, 1x50ml D29mm RB, 2/pk
14.2
-0.6
30314846
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.7
30314847
Rack, 1x50ml D29mm Rim, 2/pk
14.3
-0.5
30314848
Rack, 1x30ml D25mm Rim, 2/pk
13.8
-1.0
30314849
Rack, 1x25ml D25mm RB, 2/pk
14.4
-0.4
30314850
Rack, 2x15ml D17mm FA, 2/pk
14.8
0.0
30314851
Rack, 1x15ml D17mm Rim, 2/pk
14.5
-0.3
30314852
Rack, 7x15ml D17mm RB, 2/pk
14.5
-0.3
30314853
Rack, 9x5-7ml D13mm RB, 2/pk
14.2
-0.6
30314854
Rack, 10x1.5ml D11mm, 2/pk
9.9
-4.9
30314855
Rack, 9x4.5ml D12mm FB, 2/pk
14.0
-0.8
30314856
Rack, 7x5-7ml D13mm RB, 2/pk
11.6
-3.2
30314857
Rack, 7x5-7ml D13mm Vac, 2/pk
14.2
-0.6
30314858
Rack, 5x4-10ml D16mm Sar, 2/pk
14.3
-0.5
30314823
Rotor, Sw ing out,
4x 100ml, ID,
Sealable
Blank
NO Bucket/Adapter
14.6
30314860
Bucket, 100ml wo Cap, Sealable, 2/pk
14.2
-0.4
30314861
Bucket, 100ml w/ Cap, Sealable, 2/pk
14.2
-0.4
30314864
Rack, 1x100ml D40mm RB, 2/pk
13.8
-0.8
30314865
Rack, 1x 85ml D38mm RB, 2/pk
14.0
-0.6
30314866
Rack, 1x50ml D34mm RB, 2/pk
14.0
-0.6
30314867
Rack, 1x50ml D29mm RB, 2/pk
14.0
-0.6
30314868
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.5
30314869
Rack, 1x50ml D29mm Rim, 2/pk
13.9
-0.7
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.5
13.6
-4.6
-0.5
DE-46
30314870
Rack, 1x30ml D25mm Rim, 2/pk
13.1
-1.5
30314871
Rack, 1x25ml D25mm RB, 2/pk
14.0
-0.6
30314872
Rack, 1x15ml D17mm FA, 2/pk
14.1
-0.5
30314873
Rack, 2x15ml D17mm FA, 2/pk
14.1
-0.5
30314874
Rack, 2x15ml D17mm Rim, 2/pk
14.0
-0.6
30314875
Rack, 4x15ml D17mm RB, 2/pk
14.0
-0.6
30314876
Rack, 7x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314877
Rack, 5x1.5ml D11mm, 2/pk
14.0
-0.6
30314878
Rack, 3x10ml D17mm Sar, 2/pk
14.0
-0.6
30314879
Rack, 4x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314880
Rack, 4x10ml D17mm Vac, 2/pk
14.0
-0.6
30314881
Bucket, 2x50ml D29mm FA, w / Rack,
2/pk
14.6
0.0
30314882
Bucket, 3x15ml D17mm FA, w / Rack,
2/pk
14.6
0.0
30314883
Cy to, Rectangular Bucket, 2/pk
6.3
-8.3
30314824
Rotor, Sw ing out,
2x 3MTP, ID
Blank
NO Bucket/Adapter
12.0
30314891
Adapter, 48x 1.5/2.0ml D11mm, 2/pk
12.0
0.0
30314825
Rotor, Angle,
6x 85ml RB, ID, Hi
Blank
NO Bucket/Adapter
10.3
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
9.6
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
9.6
-0.7
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.0
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
9.8
-0.5
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
9.3
-1.0
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
10.3
0.0
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
9.5
-0.8
30314826
Rotor, Angle,
6x 85ml RB, ID
Blank
NO Bucket/Adapter
11.5
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.9
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
10.4
-1.1
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.9
-0.6
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
10.6
-0.9
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
10.4
-1.1
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
11.1
-0.4
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
10.4
-1.1
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-47
30314827
Rotor, Angle,
4x 85ml RB, ID, Hi
Blank
NO Bucket/Adapter
9.2
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
8.5
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
8.4
-0.8
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
8.9
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
8.6
-0.6
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
8.3
-0.9
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
7.5
-1.7
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
8.3
-0.9
30314828
Rotor, Sw ing out,
16x 50ml, ID
Blank
NO Bucket/Adapter
16.5
30304375
Rack, 4x50ml D29mm FA, 2/pk
16.5
0.0
30314583
Rack, 10x15ml D17mm FA, 2/pk
16.5
0.0
30314829
Rotor, Angle,
10x 50ml FA, ID
Blank
NO Bucket/Adapter
13.0
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
12.8
-0.2
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.4
-2.6
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
8.9
-4.1
30314830
Rotor, Angle,
6x 50ml RB/FA, ID
Blank
NO Bucket/Adapter
11.0
30130891
Adapter, 1x 30ml D26mm, 2/pk
10.7
-0.3
30130892
Adapter, 1x 16ml D18mm, 2/pk
10.3
-0.7
30130893
Adapter, 1x 15ml D17mm, 2/pk
10.6
-0.4
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.4
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
8.3
-2.7
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
6.7
-4.3
30314831
Rotor, Angle,
6x 50ml RB, ID, Hi
Blank
NO Bucket/Adapter
8.4
30130891
Adapter, 1x 30ml D26mm, 2/pk
8.2
-0.2
30130892
Adapter, 1x 16ml D18mm, 2/pk
7.9
-0.5
30130893
Adapter, 1x 15ml D17mm, 2/pk
8.0
-0.4
30314892
Adapter, 1x 15ml D17.5mm FA, 2/pk
7.7
-0.7
30314832
Rotor, Angle,
30x 15ml RB/FA, ID
Blank
NO Bucket/Adapter
12.5
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.3
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.5
-2.0
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
9.0
-3.5
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-48
30314834
Rotor, Angle,
12x 15ml RB/FA, ID
Blank
NO Bucket/Adapter
11.0
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.6
-0.4
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
9.1
-1.9
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
7.7
-3.4
30314835
Rotor, Angle,
44x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.7
-0.7
30314836
Rotor, Angle,
30x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
9.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
9.1
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
8.4
-1.0
30314838
Rotor, Angle,
24x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314839
Rotor, Angle,
12x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
6.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
6.3
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
5.6
-0.9
30314900
Adapter, 1x 1.5ml for ov er16000x g,
6/pk
6.4
-0.1
30553085
Rotor Sw ingOut
4x 750ml D99mm
ID SL
Blank
NO Bucket/Adapter
19.3
30553138
Adapter 32x 7ml D13mm RBx 2 V1
18.7
-0.6
30553135
Adapter 18x 15ml D17mm RBx 2 V1
18.8
-0.5
30553132
Adapter 8x 30ml D26mm RBx 2 V1
19.1
-0.2
30553131
Adapter 7x 50ml D29mm RBx 2 V1
19.1
-0.2
30553127
Adapter 3x 100ml D41mm RBx 2 V1
19.1
-0.2
30553125
Adapter 1x 250ml D62mm FBx 2 V1
18.9
-0.4
30553126
Adapter 1x 225/190ml D61mm FAx 2
V1
19.1
-0.2
30553128
Adapter 5x 50ml D29mm FAx 2 V1
19.1
-0.2
30553130
Adapter 7x 50ml D29mm FAx 2 V1
19.1
-0.2
30553133
Adapter 14x 15ml D17mm FAx 2 V1
19.2
-0.1
30553134
Adapter 17x 15ml D17mm FAx 2 V1
19.0
-0.3
30553140
Adapter 18x 5/10ml D16mm Vac/Sarx 2
V1
19.0
-0.3
30559377
Adapter 38x 1.1/1.4ml D8mm Sarx2 V1
18.9
-0.4
30553136
Adapter 21x 10ml D16mm RBx 2 V1
18.9
-0.4
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-49
30553139
Adapter 27x 1.8/7ml D13mm
Vac/Sarx 2 V1
18.8
-0.5
30553086
Rotor Angle
4x 500ml D69mm
ID
Blank
NO Bucket/Adapter
14.5
30564850
Adapter 1x 250ml D61.5mm FBx 2 V1
13.7
-0.8
30559417
Adapter 1x 175/225ml D61.5mm FAx 2
V1
13.4
-1.1
30559419
Adapter 1x 50ml D29.5mm FAx 2 V1
12.4
-2.1
30559416
Adapter 1x 250ml D60.5mm CBCx 2 V1
12.6
-1.9
30559421
Adapter 4x 30ml D25.5mm RBx 2 V1
13.8
-0.7
30559420
Adapter 3x 50ml D29mm RBx 2 V1
14.3
-0.2
30559422
Adapter 6x 15ml D17mm FAx 2 V1
13.8
-0.7
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
DE-50

Bitte in
Blockbuchstaben
ausfüllen!
11.7 Tabelle 7: Rüc knahmeformular / Dekontaminierungszertifikat
gen Sie dieses Formular allen cksendungen von Ausrüstungs- und Aufbauteilen bei!
Die ausgefül lte Erkrung über die Dekontamination ist eine Vorkehrung für di Annahme und die weitere Bearbeitung der Rücksendung. Wenn
keine entsprechend Eruterung beigefüg t ist, hren wir die Dekontamination auf Ihre Kosen durch.
Nachname; Familienname:
Organisation / Unternehmen:
Straße:
Postfach: Ort:
Telefon: Fax:
E-Mail:
Pos.
Crow d
De contaminate d
obje ct
Se rial numbe r
De scription /
Comme nt
1
2
3
4
Sind die oben aufgeführten Teile in Kontakt mit den folgenden Substanzen?
Gesundheitsgehrdende ssrige sungen, Puffer, Säuren, Basen:…………………….....
Ja Nein
Potentielle ansteckende Mittel: ……………………………………………………………
Ja Nein
Organische Reagenzien und sungsmittel: ……………………………………………………….. .. Ja Nein
Radioaktive Substanzen: ……………………………………………. α.. β.. γ
Ja Nein
Gesundheitsgehrdende Proteine: ………………………………………………………..
Ja Nein
DNA: ……………………………………………………………………………………… Ja Nein
Diese Substanzen haben die Ausrüstung/den Aufbau erreicht? ………………………. Ja Nein
Welche, falls ja:
Beschreibung der Maßnahmen zur Dekontamination der aufgeführten Teile:
Ich bestätige die ordnungsgeße Dekontaminationn:
Unternehmen/Abt . Ort und Datum:
Frontier
TM
Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Schizzo della mappa di FC5714 / FC5718 / FC5816 / FC5916
Figura 1
Schizzo della mappa di FC5718R
Figura 2
Schizzo della mappa di FC5816R / FC5916R
Figura 3
1 Coperchio Centrifuga
2 Finestra Rotore
3 Display
4 Etichetta Funzionamento
5 Interruttore Principale di Alimentazione
6 Connessione di Alimentazione
7 Rilascio di Emergenza
4
1
7
4
2
1
3
5
6
5
6
7
2
3
4
1
5
6
7
2
3
Etichetta di funzionamento per FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figura 4
Figura 5
1. Display LCD
2. Modello impostazione programma
3. Avvio centrifuga
4. Modello e selezione RPM/RCF
5. Modello e selezione intensità
Accelerazione/Decelerazione
6. Modello impostazione temperatura
(Solo FC5515R)
7. Modello impostazione tempo
8. Salvare le informazioni di impostazione
9. Centrifuga a rotazione lenta/veloce
10. Coperchio di rilascio
11. Arresto/Impostazioni centrifuga
12. Manopola/Quadrante di regolazione:
Modificare il numero
Display LCD
L'immagine qui di seguito illustra i singoli elementi del display LCD
Figura 6
1
2
3
4
5
6
7
8
9
10
11
12
Figura 7
Campi display:
A-1 Campo display – "RPM/RCF"
A-2 Campo display – "Acc/Dec" "Service"
A-3 Campo display – "Time/Prog"
A-4 Campo display – "Temp"
Messaggi/logo dei campi display
M1 "chiudi"
M2 "apri"
M3 "rotore"
M4 "Rotore-N°."
M5 "rpm"
M6 "rcf"
M7 "accel"
M8 "decel"
M9 "raggio"
M10 "programma"
M11 "errore"
M12 "servizio"
M13 "h m s"
M14 "temperatura"
M15 "preraffreddamento"
M14
M5
M6
M4
A-1
A-2
A-3
A-4
M2
M3
M1
M7
M8
M12
M15
M9
M11
M10
M13
Tabella Informazioni Rotore
Visualizzazi
one
numero del
rotore
Numero
d'ordine
Capacità
Compatibile
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
61
30304361
Rotor, Angle, 24x1.5/2.0ml, ID,
Sealable
20
30314820
Rotor, Swing out, 4x250ml, ID
21
30314821
Rotor, Angle, 6x250ml FB, ID
22
30314822
Rotor, Swing out, 4x100ml, ID
23
30314823
Rotor, Swing out, 4x100ml, ID,
Sealable
24
30314824
Rotor, Swing out, 2x3MTP, ID
25
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
26
30314826
Rotor, Angle, 6x85ml RB, ID
27
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
28
30314828
Rotor, Swing out, 16x50ml, ID
29
30314829
Rotor, Angle, 10x50ml FA, ID
30
30314830
Rotor, Angle, 6x50ml RB/FA, ID
31
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
32
30314832
Rotor, Angle, 30x15ml RB/FA, ID
33
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
34
30314834
Rotor, Angle, 12x15ml RB/FA, ID
35
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
36
30314836
Rotor, Angle, 30x1.5/2.0ml, ID,
Sealable
38
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
39
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
40
30314840
Rotor, Angle, 64x0.5ml, ID
41
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
85
30553085
Rotor SwingOut 4x750ml D99mm
ID SL
86
30553086
Rotor Angle 4x500ml D69mm ID
SOMMARIO
1. INTRODUZIONE ............................................................................................................................................................. 1
1.1 DESCRIZIONE E SCOPO ........................................................................................................................................................ 1
1.2 BREVE DESCRIZIONE ............................................................................................................................................................ 1
1.3 DEFINIZIONE DEI SEGNALI E DEI SIMBOLI DI AVVERTENZA ................................................................................................. 1
1.4 PRECAUZIONI DI SICUREZZA ................................................................................................................................................ 2
1.4.1 UTENTE ........................................................................................................................................................................ 2
1.4.2 ROTORE E ACCESSORI .................................................................................................................................................. 2
1.4.3 PROVVEDIMENTI PER LA VOSTRA SICUREZZA ............................................................................................................ 2
1.4.4 ESCLUDERE LE INFLUENZE AMBIENTALI SEGUENTI ..................................................................................................... 2
1.4.5 PROVVEDIMENTI PER LA SICUREZZA DURANTE IL FUNZIONAMENTO ........................................................................ 3
1.4.6 PERICOLI E PRECAUZIONI ............................................................................................................................................. 3
1.4.7 ABBREVIAZIONI UTILIZZATE NEL PRESENTE MANUALE ............................................................................................... 3
2. INSTALLAZIONE ............................................................................................................................................................. 4
2.1 DISIMBALLO ......................................................................................................................................................................... 4
2.1.1 IMBALLAGGIO DI CONSEGNA ...................................................................................................................................... 4
2.2 SELEZIONARE L'AMBIENTE .................................................................................................................................................. 4
2.3 INSTALLAZIONE .................................................................................................................................................................... 4
2.4 PRECAUZIONI DI SICUREZZA DURANTE IL FUNZIONAMENTO E GARANZIA ........................................................................ 5
3. FUNZIONAMENTO ......................................................................................................................................................... 5
3.1 MONTAGGIO E CARICAMENTO DEL ROTORE ...................................................................................................................... 5
3.1.1 INSTALLAZIONE DEI ROTORI ........................................................................................................................................ 5
3.1.2 CARICO DEI ROTORI D'ANGOLO ................................................................................................................................... 6
3.1.3 CARICO DEI ROTORI A BATTENTE ................................................................................................................................ 6
3.1.4 CARICO E SOVRACCARICO DEI ROTORI ........................................................................................................................ 7
3.1.5 RIMOZIONE DEL ROTORE ............................................................................................................................................. 7
3.2 CONTROLLO DEL COPERCHIO .............................................................................................................................................. 7
3.2.1 APERTURA DEL COPERCHIO ......................................................................................................................................... 7
3.2.2 BLOCCAGGIO DEL COPERCHIO ..................................................................................................................................... 8
3.3 PRESELEZIONE ..................................................................................................................................................................... 8
3.3.1 PRESELEZIONE DELLA VELOCITÀ / VALORE RCF ........................................................................................................... 8
3.3.2 PRESELEZIONE DEL TEMPO DI FUNZIONAMENTO ....................................................................................................... 9
3.3.3 PRESELEZIONE DELL'INTENSITÀ DEL FRENO E DELL'ACCELERAZIONE ........................................................................ 9
3.3.4 PRESELEZIONE DELLA TEMPERATURA (SOLO MODELLI REFRIGERATI) ...................................................................... 10
3.3.5 PRE-RAFFREDDAMENTO (SOLO MODELLI REFRIGERATI) .......................................................................................... 10
3.4 CORREZIONE DEL RAGGIO ................................................................................................................................................. 10
3.5 PROGRAMMA .................................................................................................................................................................... 11
3.5.1 SALVATAGGIO DI PROGRAMMI ................................................................................................................................. 11
3.5.2 RICHIAMO DEI PROGRAMMI SALVATI ....................................................................................................................... 12
3.5.3 LASCIARE LA MODALITÀ PROGRAMMA ..................................................................................................................... 12
3.6 AVVIAMENTO E ARRESTO DELLA CENTRIFUGA ................................................................................................................. 12
3.6.1 AVVIARE LA CENTRIFUGA .......................................................................................................................................... 12
3.6.2 IL TASTO "STOP" ........................................................................................................................................................ 13
3.7 RILEVAMENTO DI SQUILIBRI .............................................................................................................................................. 13
4. IMPOSTAZIONE ........................................................................................................................................................... 14
4.1 REGOLAZIONI DI BASE ....................................................................................................................................................... 14
4.1.1 ACCESSO ALLA MODALITÀ "DATI DI FUNZIONAMENTO" .......................................................................................... 14
4.1.2 INDICAZIONE DELLA TEMPERATURA ......................................................................................................................... 14
4.1.3 ACCENSIONE/SPEGNIMENTO SEGNALE ..................................................................................................................... 15
4.1.4 PRE-SELEZIONE DEL VOLUME DEL SEGNALE ACUSTICO ............................................................................................ 15
4.1.5 SELEZIONE DELLA SUONERIA DEL SEGNALE ACUSTICO "FINE CICLO" ...................................................................... 16
4.1.6 ACCENSIONE/SPEGNIMENTO DEL SUONO TASTIERA ................................................................................................ 16
4.1.7 RICHIAMO DEI DATI DI FUNZIONAMENTO ................................................................................................................ 17
5. MANUTENZIONE ......................................................................................................................................................... 17
5.1 MANUTENZIONE E PULIZIA ................................................................................................................................................ 17
5.1.1 GENERALITÀ .............................................................................................................................................................. 17
5.1.2 PULIZIA E DISINFEZIONE DEL GRUPPO ...................................................................................................................... 18
5.1.3 PULIZIA E DISINFEZIONE DEL ROTORE ....................................................................................................................... 18
5.1.4 DISINFEZIONE DEI ROTORI IN ALLUMINIO ................................................................................................................. 18
5.1.5 DISINFEZIONE DEI ROTORI IN POLIPROPILENE .......................................................................................................... 19
5.1.6 ROTTURA DEL VETRO ................................................................................................................................................. 19
5.2 DURATA IN VITA DI ROTORI, CONTENITORI E ACCESSORI ................................................................................................. 19
6. RICERCA GUASTI .......................................................................................................................................................... 20
6.1 MESSAGGIO DI ERRORE: CAUSA/SOLUZIONE .................................................................................................................... 20
6.2 ESAME DEI POSSIBILI MESSAGGI DI ERRORE E DELLE RELATIVE SOLUZIONI ..................................................................... 20
6.2.1 RILASCIO DEL COPERCHIO DURANTE UNA CADUTA DI TENSIONE (RILASCIO DEL COPERCHIO DI EMERGENZA)...... 20
6.2.2 DESCRIZIONE DEL SISTEMA DEI MESSAGGI DI ERRORE ............................................................................................. 21
7. RICEZIONE DELLE CENTRIFUGHE PER RIPARAZIONE ..................................................................................................... 21
8. TRASPORTO, STOCCAGGIO E SMALTIMENTO ............................................................................................................... 22
8.1 TRASPORTO ....................................................................................................................................................................... 22
8.2 STOCCAGGIO ..................................................................................................................................................................... 22
8.3 TRASPORTO, INSTALLAZIONE E SMALTIMENTO DELLA CENTRIFUGA FC5714/FC5718/FC5718R/FC5816/FC5816R ....... 22
8.3.1 TRASPORTO ................................................................................................................................................................ 22
8.3.2 INSTALLAZIONE .......................................................................................................................................................... 22
8.3.3 IMBALLAGGIO ............................................................................................................................................................ 23
8.3.4 TRASFERIMENTO DEL DISPOSITIVO ........................................................................................................................... 23
9. DATI TECNICI ............................................................................................................................................................... 24
9.1 SPECIFICHE ......................................................................................................................................................................... 24
9.1.1 CENTRIFUGA FC5714................................................................................................................................................. 24
9.1.2 CENTRIFUGA FC571................................................................................................................................................... 25
9.1.3 CENTRIFUGA FC5718R ............................................................................................................................................... 26
9.1.4 CENTRIFUGA FC5816 ................................................................................................................................................. 27
9.1.5 CENTRIFUGA FC5816R ............................................................................................................................................... 28
9.2 DISEGNI E DIMENSIONI ...................................................................................................................................................... 29
10. CONFORMITA .............................................................................................................................................................. 31
11. APPENDICE .................................................................................................................................................................. 32
11.1 TABELLA 1: PESO NETTO AMMESSO ................................................................................................................................ 33
11.2 TABELLA 2: TEMPERATURE P BASSE ALLA MAX. VELOCITÀ (SOLO MODELLI REFRIGERATI) ....................................... 34
11.3 TABELLA 3: VELOCITÀ MAX. E VALORI RCF PER ROTORI AMMISSIBILI ............................................................................ 35
11.4 TABELLA 4: TEMPI DI ACCELERAZIONE E DECELERAZIONE .............................................................................................. 37
11.5 TABELLA 5: MESSAGGI DI ERRORE ................................................................................................................................... 41
11.6 TABELLA 6: SPECIFICHE CORREZIONE RAGGIO E ADATTATORE ...................................................................................... 42
11.7 TABELLA 7: MODULO DI RISANAMENTO/CERTIFICATO DI DECONTAMINAZIONE .......................................................... 46
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-1
1. INTRODUZIONE
1.1 Descrizione e Scopo
Grazie per aver scelto questo prodotto OHAUS
Tutti i simboli indicano istruzioni di sicurezza e punti di situazioni potenzialmente pericolosi. Vi invitiamo a leggere
interamente il manuale prima di utilizzare Frontier
TM
MultiPro al fine di evitare
funzionamenti errati.
Le centrifughe Frontier
TM
MultiPro sono state progettate per la separazione di
materiali o miscele con differenti densità.
Le centrifughe OHAUS sono intese esclusivamente per utilizzo all'interno e da parte di personale qualificato.
1.2 Breve descrizione
I modelli FC5714/FC5718/FC5816/FC5916 sono centrifughe universali non refrigerate.
I modelli FC5718R/FC5816R/FC5916R sono centrifughe universali refrigerate.
Tutti e tre i modelli sono offerti in due variazioni di tensione, 230V o 120V.
Le centrifughe possono essere utilizzati con rotori a battenti e ad angolo
Tutti i parametri sono accessibili mediante pulsanti e selezionati mediante il regolatore centrale. Tutti i valori
preselezionati e attuali saranno mostrati in modo permanente sul display LCD.
La centrifuga è alimentata da un motore a induzione che non richiede manutenzione.
I dati tecnici dettagliati sono contenuti nella sezione "Dati Tecnici".
1.3 Definizione dei Segnali e dei Simboli di Avvertenza
Le avvertenze di sicurezza son contrassegnate con diciture e simboli di avvertenza Essi indicano problemi
riguardanti la sicurezza e le relative avvertenze. La mancata attenzione alle avvertenze di sicurezza può provocare
lesioni personali, danni allo strumento, malfunzionamenti e falsi risultati.
Il grado di pericolosità è una parte dell'avvertenza di sicurezza e distingue l'uno dall'altro i possibili risultati della
mancata osservanza .
Diciture
PERICOLO
Se non evitato, provocherà lesioni gravi o morte.
AVVERTENZA
Per una situazione pericolosa con rischio medio., che può provocare lesioni o more se
non evitata.
PRECAUZIONE
Per una situazione pericolosa con rischio ridotto, che può produrre danni alla proprietà
oppure perdita di dati o lesione se non evitata.
ATTENZIONE
Per informazioni importanti riguardanti il prodotto Se non evitate, possono provocare danni
all'apparecchiatura
NOTA
Per informazioni utili riguardanti il prodotto
Simboli di Avvertenza
Pericolo Generico Pericolo di Scossa Elettrica
Corrente Alternata Rischio Biologico
Esplosione Schiacciamento
IT-2 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Avvertenza e segnali di informazione sulla superficie della centrifuga
Occorre utilizzare sempre quattro supporti sui rotori a battente a quattro punti,
altrimenti la centrifuga subirà dei danni. Detti danni non saranno coperti dalla garanzia
sul prodotto.
Attenzione! Verificare il serraggio del dado del rotore prima di ogni funzionamento.
Staccare la spina di alimentazione di rete prima di aprire la sede o il rilascio
di emergenza.
Senso di rotazione - rotazione oraria per la motorizzazione del rotore
Riferimenti per il carico dei rotori
1.4 Precauzioni di sicurezza
1.4.1 Utente
Le centrifughe OHAUS sono intese esclusivamente per utilizzo all'interno e da parte di personale qualificato. Il
dispositivo può essere utilizzato solo da personale specializzato e formato. Detto personale deve aver letto
attentamente il manuale di funzionamento e avere familiarità con il funzionamento del dispositivo.
1.4.2 Rotore e accessori
Si devono utilizzare solo rotori e accessori originali OHAUS. Ogni altro utilizzo o utilizzo inteso deve essere
considerato inappropriato. OHAUS non è responsabile per danni derivanti dall'utilizzo inappropriato.
ATTENZIONE:
Leggere tutte le avvertenze di sicurezza prima di installare, connettere o eseguire manutenzione a questa
apparecchiatura. La mancata osservanza di dette avvertenze può causare lesioni personali e/o danni
alla proprietà. Conservare le istruzioni per riferimenti futuri.
1.4.3 Provvedimenti per la vostra sicurezza
AVVERTENZA: Non lavorare mai in un ambiente soggetto a pericolo di esplosione! La sede dello
strumento non è a tenuta di gas. (pericolo di esplosione dovuto alla formazione di scintille, corrosione
provocata dall'ingresso di gas)
AVVERTENZA: Quando si utilizzano prodotti chimici e solventi, conformarsi alle istruzioni del produttore e
alle regole generali
AVVERTENZA: Questa centrifuga non è sigillata. Utilizzare gli appropriati provvedimenti di sicurezza
quando si utilizza la centrifuga per campioni infettivi e patogeni, Quando si maneggiano questi campioni,
seguire le appropriate precauzioni di sicurezza.
1.4.4 Escludere le influenze ambientali seguenti
Forti vibrazioni
Luce solare diretta
Umidità atmosferica superiore all'80%
Presenza di gas corrosivi
Temperature inferiori a 2 °C e superiori a 35 °C
Forti campi elettrici o magnetici:
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-3
ATTENZIONE:
Pericolo di scosse elettriche all'interno della sede La sede deve sempre essere aperta solo da
personale autorizzato e qualificato prima di aprire, scollegare tutte le connessioni di alimentazione dal
gruppo.
1.4.5 Provvedimenti per la sicurezza durante il funzionamento
Non svitare le due metà della sede.
Asciugare immediatamente eventuali versamenti di liquidi! Lo strumento non è a tenuta stagna.
Verificare che la tensione in ingresso all'apparecchiatura e il tipo di spina siano compatibili con l'alimentazione
locale.
Connettere il cavo di alimentazione a una presa di alimentazione adeguatamente messa a terra.
Utilizzare solo un cavo di alimentazione con una classificazione superiore alle specifiche dell'etichetta
dell'apparecchiatura.
Non posizionare l'apparecchiatura in modo che sia difficile scollegare il cavo di alimentazione dalla presa di
alimentazione.
Accertarsi che il cavo di alimentazione non costituisca un potenziale ostacolo o rischio di inciampo.
L'apparecchiatura è per solo utilizzo all'interno. Utilizzare l'apparecchiatura solo in ambienti asciutti.
Utilizzare solo accessori del tipo approvato.
Azionare l'apparecchiatura solo in condizioni ambientali specificate nelle presenti istruzioni.
Disconnettere l'apparecchiatura dall'alimentazione durante la pulizia.
Non mettere l'apparecchiatura in funzione in ambienti pericolosi o instabili.
La manutenzione deve essere esclusivamente eseguita da personale autorizzato.
1.4.6 Pericoli e precauzioni
Al fine di proteggere le persone e l'ambiente, si devono osservare le precauzioni seguenti:
Durante la centrifugazione, la presenza di persone e il deposito di materiali pericolosi sono proibiti nel raggio di
30 cm dalla centrifuga conformemente alla normativa EN 61010-2-020.20.
Non centrifugare mai materiali tossici o patogeni senza le adeguate precauzioni di sicurezza, ovvero la
centrifugazione di contenitori/tubi con chiusura emetica mancante o difettosa è strettamente proibita. L'utente
ha l'obbligo di eseguire le adeguate procedure di disinfezione nel caso in cui sostanze pericolose abbiano
contaminato la centrifuga e/o i suoi accessori. Quando si centrifugano sostanze infettive, prestare sempre
attenzione alle precauzioni generali di laboratorio. Se necessario, mettersi in contatto con il vostro responsabile
della sicurezza!
E' proibito utilizzare la centrifuga con rotori differenti da quelli elencati per questo gruppo.
Non aprire in nessun caso il coperchio della centrifuga quando il rotore ruota ancora una velocità > 2m/s
1.4.7 Abbreviazioni utilizzate nel presente manuale
Simbolo/Abbreviazione
Unità
Descrizione
giri/minuto
[min
-1
]
rpm
giri/minuto
RCF
[x g]
forza centrifuga relativa
PCR
Reazione a catena polimerasi
PP
-
Polipropilene
PC
-
Policarbonato
accel
-
accelerazione
decel
-
decelerazione
prog
-
programma
FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R non sono a prova di esplosione e percnon
devono essere messe in funzione in zone o ambienti in cui vi sia rischio di esplosione. La centrifugazione di
sostanze infiammabili, esplosive, radioattive o tali da reagire chimicamente con energia elevata è strettamente
proibita. La decisione finale sui rischi associati all'utilizzo di tali sostanze è di responsabilità dell'utente della
centrifuga.
IT-4 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
2. INSTALLAZIONE
2.1 Disimballo
Rimuovere attentamente la vostra centrifuga e ogni componente dall'imballo. I componenti inclusi variano secondo
il modello di centrifuga (vedere tabella qui di seguito). Conservare l'imballaggio per garantire lo stoccaggio e il
trasporto in condizioni di sicurezza. Il manuale di istruzioni deve sempre essere conservato con la centrifuga!
Gli accessori del/dei rotore/i saranno imballati separatamente
AVVERTENZA: Pericolo da sollevamento Vi è rischio di lesioni si sollevata da una sola
persona Utilizzare un dispositivo di sollevamento meccanico oppure procedure di
sollevamento in equipe quando si solleva o si sposta l'apparecchiatura.
Fare riferimento alla sezione 8.3 per i particolari su come sollevare per estrarre dall'imballaggio.
2.1.1 Imballaggio di consegna
2.2 Selezionare l'Ambiente
NOTA!
Evitare vibrazioni eccessive, fonti di calore, correnti d'aria o rapidi cambiamenti di temperatura.
La centrifuga deve essere installata su una superficie uniforme, salda e livellata, se possibile su un
armadio/tavole da laboratorio o su una qualche altra superficie salda prova di vibrazioni.
Durante la centrifugazione, la centrifuga deve essere posta di modo che vi sia uno spazio minimo di 30 cm su
ogni lato del gruppo, conformemente alla normativa EN 61010-2-020.
Non posizionare la centrifuga in prossimità di una finestra o di un riscaldatore, ove potrebbe essere esposta a
calore eccessivo, poiché le prestazioni del gruppo si basano su una temperatura ambiente di 23°C.
2.3 Installazione
Seguire le fasi seguenti:
Verificare che l'alimentazione corrisponda a quella specificata sull'etichetta dei valoro nominali del costruttore,
che è posizionata sul pannello posteriore.
Per FC5718R e FC5816R, la linea di alimentazione deve essere protetta da un ruttore di circuito di valore 16
A (tipo K).
In caso di emergenza, vi deve essere un interruttore di emergenza installato all’esterno della stanza al fine di
disconnettere l'alimentazione dal gruppo.
Connettere la centrifuga a una presa di alimentazione messa a terra.
Connettere la centrifuga alla rete (La presa per il cavo di alimentazione deve essere facilmente raggiungibile
per essere disconnessa)
Accendere lo strumento utilizzando l'interruttore di alimentazione di rete.
Aprire il coperchio utilizzando il pulsante Door Open
Rimuovere il dispositivo di fissaggio del motore durante il trasporto.
Quantità
Descrizion e
1
Centrifuga
FC5714/F C5718/F C5718R/FC5816/F C5816R
/FC5916/FC5916R
1
Cavo di alimentazione
1
Scheda Garanzia
1
Manuale di Istruzioni Guida Rapida
1
Chiave
Rotor e
Per FC5714/FC5718/FC5816, la linea di alimentazione deve essere protetta da un ruttore di circuito di valore
10 A (tipo K).
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-5
2.4 Precauzioni di sicurezza durante il funzionamento e garanzia
Non azionare la centrifuga nel caso in cui non sia installata correttamente
Non Sporgersi sulla centrifuga durante il funzionamento
Non rimanere nel raggio di 30 cm dall''area di ingombro della macchina più a lungo del tempo necessario per
ragioni operative.
Non porre materiali potenzialmente pericolosi nel raggio di 30 cm dalla zona di ingombro.
Non azionare la centrifuga quando smontata (es: senza sede).
Non azionare la centrifuga se i componenti meccanici o elettrici sono stati manipolati.
Non utilizzare accessori quali rotori e contenitori che non siano esclusivamente approvati da OHAUS
Corporation, salvo tubi centrifuga di vetro o plastica disponibili in commercio.
Non trattare materiali estremamente corrosivi, che possono danneggiare o indebolire i materiali.
Non azionare la centrifuga con rotori o recipienti che indichino eventuali segni di corrosione o danni meccanici.
Il costruttore è responsabile per la sicurezza e l'affidabilità della centrifuga solo se:
1) Il gruppo è fatto funzionare conformemente al presente manuale di istruzioni.
2) Modifiche, riparazioni o altre regolazioni sono eseguite da personale autorizzato OHAUS e l'impianto
elettrico è conforme al relativo codice elettrico.
NOTA!
Garanzia
La centrifuga è stata sottoposta ad approfondito collaudo e controllo qualità. Nell'improbabile caso di difetti di
fabbricazione, la centrifuga e i rotori sono coperti da garanzia. Detta garanzia decade in caso di errata
manipolazione, danni e negligenze e, inoltre nel caso di utilizzo di parti di ricambio e/o accessori inappropriati,
oppure modifiche non autorizzate al gruppo.
I diritti alle modifiche tecniche sono riservate al costruttore per quanto attiene miglioramenti tecnici.
3. FUNZIONAMENTO
3.1 Montaggio e caricamento del rotore
3.1.1 Installazione dei rotori
Pulire l'albero motore e la bussola di chiusura con un panno pulito e privo di grasso. Posizionare il rotore sull'albero
motore. (Vedere figura qui di seguito). Prestare attenzione che il rotore sia completamente installato sull'albero
motore.
Figura 8
Figura 9 Figura 10
Dado per
Rotore
Utensile per
rotore con dado
Utensile per rotore
senza dado
Coperchio a
scatto
Coperchio
avvitabile
Albero motore e camera
IT-6 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Tenere il rotore con una mano e fissarlo all'albero ruotando il dado di fissaggio in senso orario. Serrare il dado di
fissaggio con la chiave rotore fornita (Vedere figure 9-10)
Forniremo un utensile per fissare il rotore alla centrifuga, l'utensile per il fissaggio del dado al rotore sarà fornito
con il rotore
ATTENZIONE!
Verificare che la vite di fissaggio sia adeguatamente installata prima di ogni utilizzo. (Vedere figura 9-
10) Non azionare la centrifuga con rotori o recipienti che indichino eventuali segni di corrosione o
danni meccanici. Non operare con sostanze molto corrosive, che potrebbero danneggiare il rotore, i
recipienti e i materiali. In caso di domande, si prega di prendere contatto con il costruttore!
3.1.2 Carico dei rotori d'angolo
I rotori devono essere caricati simmetricamente e con lo stesso peso (Vedere figura qui di seguito). L'adattatore
può essere caricato solo con recipienti approvati. La differenza di peso fra i recipienti pieni deve essere la minima
possibile. Consigliamo perciò di pesarli con una bilancia. Ciò consente di ridurre l'usura della motorizzazione e il
rumore durante il funzionamento.
Il carico massimo per foro è dichiarato su ogni rotore.
Figura 11ERRATO Figura 12 CORRETTO (6 tubi)
3.1.3 Carico dei rotori a battente
Il carico dei recipienti deve essere eseguito conformemente alla figura qui di seguito.
E' solo consentito operare, per esempio, con un rotore da 4 posizioni con solo 2 contenitori caricati. I contenitori
caricati devono però essere opposti gli uni agli altri. Accertarsi che i contenitori scaricati siano messi all'interno del
rotore (vedere qui sotto).
In linea di principio, i rotori a battente non possono essere messi in funzione fino a che tutti i recipienti o le
rastrelliere siano posizionati nel rotore.
Le viti sul rotore devono essere lubrificate con olio "High TEF". I tubi campione devono essere riempiti
uniformemente a occhio nei fori o nelle rastrelliere dei tubi. La differenza di peso dei recipienti caricati non deve
essere superiore a 1,0 g circa.
ATTENZIONE!
I rotori a battente devono essere messi in funzione solo se tutte le sedi sono riempite con quattro
recipienti o quattro attrezzi: non mischiare recipienti e attrezzi!
ATTENZIONE!
Non azionare la centrifuga con rotori o recipienti che indichino eventuali segni di corrosione o
danni meccanici.
Non operare con sostanze molto corrosive, che potrebbero danneggiare il rotore, e i recipienti.
In caso di domande, si prega di prendere contatto con il costruttore!
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-7
Figura 13 ERRATO Figura 14 CORRETTO
3.1.4 Carico e sovraccarico dei rotori
Tutti i rotori approvati sono elencati con la loro velocità max. e peso max. di riempimento in "tabella 2
peso netto ammissibile" (Vedere APPENDICE).
Il carico max. ammesso per un rotore, che è determinato dal costruttore, e anche la massima velocità ammessa
per questo rotore (Vedere etichetta sul rotore), non devono essere ecceduti. I liquidi con cui si carica il rotore
devono avere una densità omogenea max. di 1/2 g/l o inferiore quando il rotore funziona alla velocità max.
Per centrifugare liquidi con una densità più elevata, occorre ridurre la velocità conformemente alla formula
seguente:
Velocità ridotta n
red
=
1,2
𝑖𝑔𝑒𝑟 𝑑𝑒𝑛𝑠𝑖𝑡𝑦
x velocità max. (n
max
) del rotore
Esempio:
n
red
=
1,2
1,7
x 4.000 = 3.360 rpm
In caso di domande, si prega di prendere contatto con il costruttore!
Al fine di determinare la forza centrifuga relativa (RCF/forza di accelerazione) per un adattatore specifico, è
possibile calcolare secondo DIN 58 970 utilizzando la formula allegata:
RCF = 1.117862*10
5
*n
2
*r
max
n: giri/minuto (RPM)
r
max
: raggio di centrifuga max in cm utilizzando il fondo dei tubi
3.1.5 Rimozione del rotore
Svitare completamente il dado di fissaggio del rotore (vite sul punto rigido) e sollevare verticalmente il rotore fuori
dalla centrifuga. (Vedere figure 9 e 10)
3.2 Controllo del coperchio
3.2.1 Apertura del coperchio
Dopo il funzionamento, quando il coperchio della centrifuga è chiuso, appare la parola "close" (M1) sul display
"RPM | RCF" (A-1). Inoltre, se vi è un rotore nella centrifuga, appare la parola "rotor"(M3) e anche il numero di
codice del rispettivo rotore che si trova all'interno del sistema centrifugo "71" (M4). Se nella centrifuga non vi è
alcun rotore, la parola "rotor" (M3) lampeggia e appare la parola supplementare "no" (M4). Premendo il tasto
"Door Open" (7), si può rilasciare il coperchio della centrifuga. Non appena il coperchio elettromagnetico è
completamente rilasciato, appare la parola "open" (M2). Adesso è possibile aprire il coperchio della centrifuga
Fare riferimento alla figura 15 qui di seguito per riferimento
IT-8 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Figura 15
Durante il funzionamento, è possibile richiamare in qualunque momento il tipo di rotore premendo il tasto "Door
Open" (7).
3.2.2 Bloccaggio del coperchio
Il coperchio deve solo essere abbassato leggermente Un bloccaggio elettromagnetico del coperchio fa chiudere il
coperchio stesso, e allo stesso tempo scompare la parola "open" (M2) (fare riferimento alla figura 15).
Quale segnale che la centrifuga è pronta per essere avviata, il display "RPM | RCF" (A-1) visualizza la parola
"close" (M1) . Allo stesso tempo, si visualizza la parola "rotor" (M3), come il numero di codice del rotore che si
trova nell'impianto della centrifuga, "no 71" (M4). Con questo, tutti i dati specifici del rotore, quali, per esempio. la
velocità max. l'accelerazione, ecc. sono adottati.
ATTENZIONE:
Non lasciare impigliare le dita fra il coperchio e il dispositivo o il meccanismo di bloccaggio quando si
chiude il coperchio.
3.3 Preselezione
3.3.1 Preselezione della velocità / valore RCF
Questa preselezione è attivata mediante il tasto "RPM | RCF" (4) (fare riferimento alla figura 16 qui di seguito).
Premendo una volta il tasto, la parola "rpm" (M5) lampeggia. Premendo due volte il tasto, è possibile selezionare
la preselezione delle forze centrifughe. Poi compare la parola "rcf" (M6) lampeggiante. Si possono impostare i
valori desiderati mediante la manopola di regolazione (1). Nel display (A-1), il valore regolato è visualizzato in
permanenza, prima, durante e dopo il funzionamento.
Figura 16
M5
A-1
4
5
M6
M3
A-1
M2 M4 M1
7
1
1
Fino a che non vi è rotore inserito, la velocità è regolabile fra 200 giri/minuto e la rotazione massima della
centrifuga.
Se vi è un rotore nella centrifuga, la velocità può essere preselezionata solo fino alla massima rotazione ammessa
per quel rotore. Lo stesso vale con la preselezione del valore
RCF.
Vedere "Tabella 3: velocità max. e valori RCF per il rotore ammesso " (Vedi APPENDICE). Tutti i valori
importanti sono elencati lì.
L'intervallo di impostazione è compreso tra la forza centrifuga a 200 rpm e la forza centrifuga massima consentita del
rotore.
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-9
ATTENZIONE:
Si prega anche di verificare con il costruttore il numero ammesso di rotazioni delle vostre provette.
3.3.2 Preselezione del tempo di funzionamento
Il tempo di funzionamento può essere preselezionato in tre gamme differenti, da 10 secondi fino a 99 ore e 59
minuti.
1. Gamma da 10 secondi fino a 59 minuti e 50 secondi in passi di 10 secondi
2. Gamma da 1 secondi fino a 99 minuti e 59 minuti in passi di 1 minuto
3. Il funzionamento continuo "cont", può essere interrotto con il tasto "Stop"(10) (fare riferimento alla figura17).
Il tempo di funzionamento può essere preselezionato con il coperchio aperto o chiuso.
Per attivare le impostazioni del tempo di funzionamento, premere il tasto "Time" (6).
Nel display "Time/Prog" (A-3), l'indicazione "m : s" o "h : m" lampeggia, secondo l'impostazione precedente.
Per impostare il valore desiderato, utilizzare la manopola di regolazione (1). Dopo aver superato 59 minuti e 50
secondi, l'indicazione cambia automaticamente a "h : m". Dopo aver superato 99 ore 59 minuti, la parola "cont"
appare nel display "Time/Prog" (A-3). Il funzionamento continuo può essere interrotto solo mediante il tasto
"Stop"(10). Il conteggio alla rovescia del tempo incomincia non appena si raggiunge la velocità impostata.
Il display indica sempre il tempo di funzionamento rimanente. (Vedere figura 17)
Figura 17
3.3.3 Preselezione dell'intensità del freno e dell'accelerazione
Questa funzione è attivata mediante il tasto "Accel/Decel" (5) (fare riferimento alla figura 18).
Premendo il tasto una volta, la parola "accel" (M7) lampeggia nel display "Acc/Dec" (A-2). L'accelerazione
desiderata può essere preselezionata mediante la manopola di regolazione (1). Il valore 0 è equivalente al più
basso e il valore 9 è l'accelerazione più elevata.
Premendo due volte il tasto "Accel/Decel" (5), il display "Acc/Dec" (A-2), visualizza la parola "decel"(M8).
Adesso è possibile preselezionare l'intensità di frenata desiderata mediante la manopola di regolazione (1). Il
valore 9 è equivalente al più breve e il valore 0 è il tempo di frenata più lungo.
Vedere "tabella 5: tempi di accelerazione e decelerazione " (APPENDICE). Qui sono indicati i tempi di
accelerazione e decelerazione per le fasi di accelerazione e decelerazione da 0 a 9 per i rotori ammessi.
Figura 18
6
A-3
1
10
M7
M8
5
A-2
1
IT-10 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R /FC5916/FC5916R
3.3.4 Preselezione della temperatura (Solo Modelli Refrigerati)
Questa funzione è attivata dal tasto "Temp/Setup" (13). Dopo aver premuto questo tasto, nel display
"Time/Prog" lampeggia l'indicazione "°C" (A-4). Mediante la manopola di regolazione (1) è possibile
preselezionare la temperatura di test in passi da 1°C in una gamma da -20°C fino a +40°C.
Questo valore è indicato in modo permanente nel display (figura 19), prima durante e dopo il funzionamento.
Si prega di notare le rispettive temperature più basse dei rotori alla velocità max.!
Figura 19
3.3.5 Pre-raffreddamento (Solo Modelli Refrigerati)
Se i campioni sono sensibili alla temperatura, è utile preraffreddare la centrifuga, il rotore e in ultimo i contenitori
alla temperatura di funzionamento necessaria. Perciò, inserire il rotore desiderato e preimpostare la rispettiva
temperatura. Si avvia il ciclo premendo simultaneamente i tasti "Temp/Setup" (13) (fare riferimento alla figura 20)
e "Time" (6). Durante il funzionamento, il gruppo sceglie automaticamente una velocità di rotazione equivalente al
20% della velocità di rotazione del rispettivo rotore. Dopo aver raggiunto la temperatura preimpostata, è possibile
lasciare il ciclo di preraffreddamento con il tasto "Stop" (10).
Secondo il rotore inserito, il preraffreddamento va da circa 10 a 20 minuti.
Figura 20
3.4 Correzione del raggio
Se si utilizzano adattatori o riduttori, è possibile modificare il raggio centrifugo del rispettivo rotore. Nel qual caso, il
raggio può essere corretto manualmente. Procedere come segue:
Chiudere il coperchio, quindi premere allo stesso tempo il tasto "Time" (6) (fare riferimento alla figura 21) e il tasto
"Prog/Setup" (11) e tenerli premuti.
Il display "Time/Prog" (A-3) visualizza la parola "radius" (M9). Regolando la manopola (1), è possibile
preselezionare la rispettiva correzione del raggio (Vedere tabella 7, APPENDICE) in passi di 0.1 cm. Non appena
si è selezionata una correzione del raggio, compare la parola "radius" (M9). Questa parola rimarrà visibile fino a
che non si rimette la correzione del raggio a 0.
A-4
1
13
A-4
13
6
10
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-11
Figura 21
3.5 Programma
3.5.1 Salvataggio di programmi
Si possono salvare fino a 99 cicli con tutti i relativi parametri, compresi i rotori utilizzati. Si può utilizzare qualunque
numero di programma libero e richiamarlo.
Porre nella centrifuga il rotore necessario. Premendo il tasto "Prog/Setup" (11) sul display "Time/Prog" (A-3),
appare la parola "programm" . E' possibile scegliere il numero di programma desiderato mediante la manopola di
regolazione (1).
Se un numero di programma è già occupato, sul display "RPM | RCF" (A-1), appariranno le parole "rotor" (M3) a
"xx" (M4). Nel caso di numeri di programma liberi, appare 0
Figura 22
Chiudere il coperchio della centrifuga Adesso procedere come descritto in precedenza per impostare tutti i
parametri di ciclo importanti. Se il coperchio non è chiuso quando si salva il programma, sul display "RPM/RCF"
(A-1), lampeggiano alternativamente le parole "FirSt" e "CLOSE Lid" (Vedere figura 23). Se si vuole avviare il
ciclo senza salvare il programma, sul display "RPM/RCF" (A-1) lampeggiano alternativamente le parole "First" e
"PrESS StoreE" (Vedere figura 24) .
Figura 23 Figura 24
11
A-3
A-1
M3
M4
M10
1
12
1
11
A-3
M9
6
IT-12 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Per adottare i dati, premere il tasto "Store" (12) (fare riferimento alle figure 23 e 24) per circa 1 secondo. Se il
programma è salvato correttamente, la parola "Store" appare sul display "RPM/RCF" (A-1). Per effetto di questo,
la parola "programm" (M10) scompare.
Non appena si rilascia il tasto "Store" (12), ricompare la parola "programm xx" (M10) , ove (xx) sta per la
posizione scelta per il programma.
Se tutti i numeri di programma sono occupati, si può prendere un vecchio numero che non sia più necessario e
semplicemente digitare i nuovi parametri.
3.5.2 Richiamo dei programmi salvati
Pre richiamare i programmi salvati, premere il tasto "Prog/Setup" (11) (fare riferimento alla figura 25) quando il
coperchio è già chiuso. Sul display "Time/Prog" (A-3), appare la parola "programm –"(M10). Il numero di
programma desiderato può essere preselezionato mediante la manopola di regolazione (1).
Nei rispettivi display appariranno i valori salvati per quel programma.
Se all'interno della centrifuga si trova un rotore errato per il programma preselezionato, sul display "RPM | RCF"
(A-1), lampeggerà la parola "rotor" (M3) . Nello stesso tempo, lampeggiano a turno la parola "FALSE" e il
numero di rotore salvato " xx" (M4).
Figura 25
3.5.3 Lasciare la modalità programma
Per lasciare la modalità programma, premere semplicemente il tasto "Prog/Setup" (11) (fare riferimento alla fig.
25). Sul display "Time/Prog" appare la parola "programm" .
Impostare il display a "programm--" (M10) con la manopola di regolazione (1).
3.6 Avviamento e arresto della centrifuga
3.6.1 Avviare la centrifuga
E' possibile avviare la centrifuga con il tasto "Start" (9) (rif. figura 26) oppure con il tasto "Quick Spin" (8).
Con il tasto "Start" (9), si possono avviare cicli salvati oppure cicli con i parametri preselezionati manualmente.
Quando è finito il rispettivo tempo ciclo preselezionato, la centrifuga si arresterà automaticamente.
Con il tasto "Quick Spin" (8), si possono avviare cicli, che dureranno appena qualche secondo.
Premendo il tasto "Quick Spin" (8), la centrifuga accelera fino alla rotazione selezionata.
Sul display "Time/Prog" (A-3), il tempo ciclo trascorso è indicato a partire da quando si è premuto il tasto "Quick
Spin" (8).
Rilasciando il tasto "Quick Spin" (8), la centrifuga si arresta e il tempo ciclo è indicato fino all'apertura del
coperchio.
A-3
M10
A-1
M3
M4
1
11
12
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-13
Figura 26
3.6.2 Il tasto "STOP"
Mediante il tasto "Stop" (10) (Vedere figura27) è possibile interrompere il ciclo in ogni momento. Dopo aver
premuto il tasto, la centrifuga decelera all'intensità preselezionata fino a che non si arresta.
Figura 27
3.7 Rilevamento di squilibri
Nel caso in cui il rotore non sia caricato uniformemente, la motorizzazione si disconnetterà durante la
decelerazione. Il rotore decelera fino a che non si arresta.
Quando sul display "Time/Prog" (A-3) appare la parola "error” (M11) insieme al numero "01” , la differenza dei
campioni è eccessiva. Distribuire uniformemente il peso.
Caricare il rotore come descritto ai capitoli 3.1.2 e 3.1.3.
Quando sul display "Time/Prog" (A-3) appare la parola "error” insieme al numero "02" (Vedere figura 28), ciò
può essere dovuto alla ragione seguente. Difetto dell'interruttore di squilibrio.
Figura 28
10
A-3
M11
1
8
9
A-3
IT-14 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4. IMPOSTAZIONE
4.1 Regolazioni di base
4.1.1 Accesso alla modalità "Dati di funzionamento"
Quando si utilizza la centrifuga, è possibile impostare i parametri seguenti:
- Indicazione di temperatura °C o °F
- Accensione/spegnimento del segnale acustico
- Accensione/spegnimento del suono tastiera
- Pre-selezione del volume del segnale acustico
- Selezione della suoneria del segnale acustico "fine ciclo"
Mentre la centrifuga è disconnessa, premere simultaneamente i tasti "Time"(6) e "Door Open" (7) e accendere
l'interruttore principale della centrifuga. Adesso, rilasciare entrambi i tasti e - per effetto di ciò - viene eseguita una
prova del display per circa 5 secondi. Tutti gli indicatori appariranno allo stesso tempo (Vedere figura 29).
Figura 29
ATTENZIONE:
Si noti che è necessario inserire il programma descritto al punto 4.1.1 per modificare le regolazioni dei
punti 4.1.2 - 4.1.7. Dopo che si sono salvate le impostazioni, ritornare nuovamente alla modalità
normale disconnettendo la centrifuga per un breve tempo.
Tutte le impostazioni modificate devono essere confermate con il tasto "Start"(9). La parola "Store"(12) appare
sul display "RPM | RCF"(A-1) - Nelle preselezioni sono validi solo!
Figura 30
4.1.2 Indicazione della temperatura
Procedere come descritto al punto 4.1.1 per inserire questa modalità programma e poi premere il tasto
"Accel/Decel" (5). Sul display "Acc/Dec" (A-2) compare la parola "Service" . Selezionare adesso la lettera "C"
con la manopola di regolazione (1). Di conseguenza, sul display "RPM | RCF" (A-1), appaiono le parole
"CELSI/temp" . Se si preme il tasto "RPM | RCF" (4), lampeggia la parola "CELSI" e si può modificare il display
Fahrenheit "FAREN", con la manopola di regolazione(1) (Vedere figura 31). Dopo che si sono salvate le
impostazioni (Vedere 4.1.1) ritornare nuovamente alla modalità normale disconnettendo la centrifuga per un breve
tempo.
6
7
9
A-1
12
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-15
Figura 31
4.1.3 Accensione/Spegnimento segnale
Procedere come descritto al punto 4.1.1 per inserire questa modalità programma e poi premere il tasto
"Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" . Selezionare adesso la lettera "L
con la manopola di regolazione (1). Di conseguenza, compaiono le parole "On Sound" sul display "RPM | RCF"
(4). Se adesso si preme il tasto "RPM | RCF" (4), lampeggia la parola "On" e si può disattivare la suoneria, con la
manopola di regolazione(1) (Vedere figura 32).
Dopo che si sono salvate le impostazioni (Vedere 4.1.1) ritornare nuovamente alla modalità normale
disconnettendo la centrifuga per un breve tempo.
Figura 32
4.1.4 Pre-selezione del volume del segnale acustico
Procedere come descritto al punto 4.1.per inserire questa modalità programma e poi premere il tasto
"Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" . Selezionare adesso la lettera "U
con la manopola di regolazione (1). Di conseguenza, sul display "RPM | RCF" (A-1), appaiono le parole "Vol=0-
9/Sound" . Dopo aver premuto il tasto "RPM | RCF" (4), è possibile regolare il volume desiderato fra 0 (basso) e
9 (alto) con la manopola di regolazione (1) (Vedere figura 33).
Dopo che si sono salvate le impostazioni (Vedere 4.1.1) ritornare nuovamente alla modalità normale
disconnettendo la centrifuga per un breve tempo.
Figura 33
4
5
1
A-1
A-2
1
4
5
A-1
A-2
4
5
1
IT-16 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
4.1.5 Selezione della suoneria del segnale acustico "fine ciclo"
Procedere come descritto al punto 4.1.per inserire questa modalità programma e poi premere il tasto
"Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" . Selezionare adesso la lettera "G
con la manopola di regolazione (1). Di conseguenza, sul display "RPM | RCF" (A-1), appare la parola
"SonGo/Sound" . Dopo aver premuto il tasto "RPM | RCF" (4), è possibile selezionare una suoneria con la
manopola di regolazione(1). (Vedere figura 34)
Dopo che si sono salvate le impostazioni (Vedere 4.1.1), si può ritornare nuovamente alla modalità normale
disconnettendo la centrifuga per un breve tempo.
Figura 34
4.1.6 Accensione/spegnimento del suono tastiera
Procedere come descritto al punto 4.1.per inserire questa modalità programma e poi premere il tasto
"Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" . Selezionare adesso la lettera "B
con la manopola di regolazione (1). Di conseguenza, sul display "RPM | RCF" (A-1), appare la parola “ON/BEEP
. Dopo aver premuto il tasto "RPM | RCF" (4), è possibile selezionare il suono tastiera con la manopola di
regolazione(1). (Vedere figura 39).
Dopo che si sono salvate le impostazioni (Vedere 4.1.1), si può ritornare nuovamente alla modalità normale
disconnettendo la centrifuga per un breve tempo.
Figura 35
A-2
A-1
4
5
1
4
5
1
A-2
A-1
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-17
4.1.7 Richiamo dei dati di funzionamento
ATTENZIONE:
Questa operazione deve essere eseguita solo da utenti avanzati o da un tecnico di manutenzione.
I dati di funzionamento della centrifuga possono essere richiamati nella modalità "Basic
Adjustments" . Per inserire questa modalità di programma, procedere come descritto al punto 4.1.2. Premere il
tasto "Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" .
E' possibile avere accesso alle differenti informazioni mediante la manopola di regolazione (1):
A= avviamenti precedenti della centrifuga
H= ore precedenti di funzionamento
S= versione software
r= software di conversione
E= elenco dei messaggi di errore precedenti
h= tempo di funzionamento del motore
E' possibile visualizzare l'elenco degli ultimi 99 messaggi di errore premendo il tasto "RPM | RCF" (4) e scorrerlo
con la manopola di regolazione (1). I rispettivi codici di errore compaiono nel display "RPM | RCF" (A-1). Fare
riferimento a "Tabella 6: messaggio di errore " (Vedere APPENDICE).
Per ritornare ala modalità programma normale, spegnere la centrifuga per un breve tempo.
Figura 36
5. MANUTENZIONE
5.1 Manutenzione e pulizia
5.1.1 Generalità
Attenzione:
La manutenzione della centrifuga si limita al tener puliti il rotore, la camera del rotore e gli accessori del rotore
stesso, e anche a lubrificare regolarmente le viti di inserto del rotore per un rotore a battente (se disponibile).
Il lubrificante più appropriato è l'olio High TEF.
Lubrificanti contenenti molycote e grafite non sono consentiti
Si prega di prestare particolare attenzione alle parti in alluminio anodizzato. Anche piccoli danni possono provocare
la rottura del rotore.
Nel caso in cui rotori, contenitori o rastrelliere di tubi vengano in contatto con sostanze corrosive, i punti interessati
devono essere puliti con attenzione.
Sostanze corrosive sono, per esempio: alcali, soluzioni di sapone, amine alcaline, acidi concentrati, soluzioni
contenenti metalli pesanti, solventi clorati privi di acqua, soluzioni saline, quali acqua salata, fenolo e idrocarburi
alogenati.
4
5
1
A-2
A-1
IT-18 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Pulizia – gruppi, rotori e accessori:
- Spegnere il dispositivo e disconnetterlo dall'alimentazione prima di incominciare qualunque operazione di
pulizia e disinfezione. Non versare liquidi all'interno della sede
- Non aspergere disinfettante sul dispositivo.
- Un'attenta pulizia non serve solo a disinfettare, ma anche a evitare la corrosione dovuta all'inquinamento.
- Al fine di evitare di danneggiare parti anodizzate quali rotori, piastre di riduzione, ecc. per la pulizia è possibile
utilizzare solo detergenti a pH neutro con un valore pH- da 6 a 8. Non si devono utilizzare agenti detergenti
alcalini (valore pH > 8).
- Dopo la pulizia, accertarsi che tutti gli elementi siano ben asciutti, manualmente o mediante una cassetta ad
aria riscaldata (Temperatura max.+ 50°C).
- E' necessario rivestire regolarmente gli elementi in alluminio anodizzato con olio anti-corrosione al fine di
aumentare la loro durata in vita e di ridurre la predisposizione alla corrosione.
- Per via dell'umidità o di campioni non ermeticamente chiusi, è possibile si formi della condensa. La condensa
deve essere rimossa regolarmente dalla camera del rotore con un panno soffice.
La procedura di manutenzione deve essere ripetuta ogni 10-15 funzionamenti, o almeno una
volta la settimana.
- Connettere il gruppo all'alimentazione, una volta che il macchinario sia completamente asciutto.
- Non eseguire disinfezioni con raggi UV, beta e gamma o con altre radiazioni a energia elevata.
- I rotori di metallo possono essere puliti in autoclave .
- Anche il coperchio e gli adattatori del rotore possono essere puliti in autoclave (max. 121°C, 20 min).
- Le rastrelliere dei tubi sono in polipropilene e non possono essere puliti e in autoclave a 134°C.
5.1.2 Pulizia e disinfezione del gruppo
1. Aprire il coperchio prima di spegnere il gruppo. Disconnetterlo dall'alimentazione
2. Aprire il dado del rotore ruotando in senso antiorario la chiave del rotore.
3. Rimuovere il rotore.
4. Per pulire e disinfettare il gruppo e la camera del rotore, utilizzare l'agente di pulizia citato sopra.
5. Pulire tutte le zone accessibili del dispositivo e dei suoi accessori, compreso il cavo di alimentazione, con un
panno umido.
6. Lavare a fondo le guarnizioni di gomma e la camera del rotore con acqua.
7. Cospargere le guarnizioni di gomma asciutte con glicelolo o talco al fine di prevenire che divengano fragili. Altri
componenti del gruppo, per esempio il bloccaggio del coperchio, l'albero motore e il rotore, non devono essere
lubrificati.
8. Asciugare l'albero motore con un panno morbido, asciutto e senza sfilacciature.
9. Verificare che il gruppo e gli accessori non abbiano subito danni.
Accertarsi che la centrifuga sia spenta e disconnettere il gruppo dall'alimentazione. Rimuovere quindi la polvere
che aderisce alle asole di ventilazione nella centrifuga mediante una spazzola soffice. Eseguire questa operazione
almeno ogni sei mesi.
5.1.3 Pulizia e disinfezione del rotore
1. Pulire e disinfettare i rotori, i coperchi del rotore e gli adattatori con l'agente di pulizia citato sopra.
2. Utilizzare uno scovolo per pulire e disinfettare gli alesaggi del rotore.
3. Risciacquare i rotori, il coperchio del rotore e l'adattatore con acqua pulita. In particolare, i fori dei rotori ad angolo.
4. Per asciugare i rotori e gli accessori, porli su un telo. Posizionare i rotori ad angolo con gli alesaggi verso il basso
5. Asciugare il cono del rotore con un panno morbido, asciutto e senza sfilacciature e verificare la presenza di
eventuali danni. Non lubrificare il cono del rotore.
5.1.4 Disinfezione dei rotori in alluminio
Nel caso in cui materiale infettivo penetri nella centrifuga, il rotore e la camera del rotore devono essere disinfettati
immediatamente dopo il funzionamento. I rotori possono essere puliti in autoclave alla temperatura max. di 121° C.
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-19
5.1.5 Disinfezione dei rotori in polipropilene
Pulizia in autoclave
Il tempo consigliato per la pulizia in autoclave da 15 a 20 min a 121°C (1 bar)
ATTENZIONE:
Il tempo di sterilizzazione di 20 minuti non deve essere ecceduto. Una sterilizzazione ripetuta
provocherà una riduzione della resistenza meccanica della materia plastica.
Prima di essere passato in autoclave, il rotore e l'adattatore in polipropilene deve essere pulito attentamente al fine
di evitare che i residui di sporcizia brucino.
Le conseguenze di alcuni residui di prodotti chimici nei materiali plastici a temperatura ambiente possono essere
trascurate. A temperature elevate, però, durante il passaggio in autoclave questi residui possono corrodere e
distruggere la plastica. Gli oggetti devono essere risciacquati a fondo con acqua distillata dopo la pulizia, ma prima
del passaggio in autoclave, I residui di qualunque liquido di pulizia possono provocare fessurazioni, sbiancamento
e macchie.
Sterilizzazione a gas
Gli adattatori, le bottiglie e i rotori possono essere sterilizzati con gas di ossido di Etilene Accertarsi di aerare gli
articoli dopo la sterilizzazione e prima di utilizzarli nuovamente.
ATTENZIONE:
Poiché durante la sterilizzazione la temperatura può salire, i rotori, gli adattatori e le bottiglie
non devono essere chiuse e devono essere completamente svitate.
Sterilizzazione chimica
Le bottiglie, gli adattatori e i rotori possono essere trattati con i normali disinfettanti chimici.
ATTENZIONE:
Prima di applicare qualunque altro metodo di pulizia o di decontaminazione non espressamente
consigliato dal costruttore, prendere contatto con il costruttore per garantire che non danneggi il
gruppo o il rotore.
5.1.6 Rottura del vetro
Con coefficienti G elevati aumenta la velocità di rottura dei tubi di vetro . I frammenti di vetro devono essere rimossi
immediatamente dal rotore, dai contenitori, dagli adattatori e dalla camera stessa del rotore. Piccoli frammenti di
vetro graffieranno e perciò danneggeranno il rivestimento protettivo di superficie di un rotore. Se frammenti di vetro
rimangono nella camera del rotore, si accumulerà polvere metallica fine per via della circolazione dell'aria. Questa
polvere metallica nera molto sottile inquinerà molto gravemente la camera del rotore, il rotore, i contenitori e i
campioni.
Se necessario, sostituire gli adattatori, i tubi e gli accessori al fine di evitare ulteriori danni. Verificare regolarmente
che gli alesaggi del rotore non contengano residui o presentino danni.
ATTENZIONE:
Verificare le specifiche del costruttore dei tubi per la centrifuga.
5.2 Durata in vita di rotori, contenitori e accessori
I rotori e il coperchio del rotore in alluminio o in acciaio inox hanno una durata di funzionamento max. di 7 anni dal
primo utilizzo.
I coperchi trasparenti del rotore in PC o PP, come anche i rotori, le rastrelliere dei tubi e gli adattatori in PP hanno
una durata in vita max. di 3 anni dal primo utilizzo.
Condizione per la durata in vita: Utilizzo appropriato, condizione priva di danni e cura consigliata.
IT-20 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
6. RICERCA GUASTI
6.1 Messaggio di errore: Causa/Soluzione
I messaggi di errore sono elencati per consentire una più rapida localizzazione degli errori.
E' possibile che la diagnosi cui si fa riferimento in questo capitolo non siano sempre quelle reali, poiché si tratta di
errori e soluzioni che avvengono solo teoricamente.
Vi preghiamo di tenerci informati di ogni tipo di errore che si verifichi e non sia fra quelli elencati nel presente
capitolo. Solo le vostre informazioni ci consentono di migliorare questo manuale di funzionamento.
Vi ringraziamo anticipatamente per il vostro sostegno.
6.2 Esame dei possibili messaggi di errore e delle relative soluzioni
6.2.1 Rilascio del coperchio durante una caduta di tensione (Rilascio del Coperchio di Emergenza)
In caso di caduta di tensione o di malfunzionamento, il coperchio della centrifuga può essere aperto al fine di
prelevare i vostri campioni.
Per FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (bloccaggio motorizzato)
Procedere come segue:
1. Spegnere la centrifuga e disconnettere il cavo di alimentazione, attendere fino a che il rotore non sia fermo
(possono essere necessari alcuni minuti)
2. Sul lato sinistro della sede della centrifuga vi è un arresto di plastica Rimuovere l'arresto e dietro di esso vi è un
dado esagonale.
3. Utilizzare la chiave di fornitura per la cassetta, porla nel foro e bloccare il dado esagonale con la chiave
(Vedere figura 37).
4. Adesso ruotare la chiave sul lato destro (senso orario) fino al limite.
ATTENZIONE:
a) Ruotare semplicemente fino al limite, non serrare il dado.
b) Adesso aprire il coperchio della centrifuga
c) Accendere nuovamente la centrifuga, per riprendere il lavoro.
Figura 37
Per 5714
Procedere nel modo seguente (vedere Figura 38):
ATTENZIONE:
Spegnere la centrifuga e disconnettere il cavo di alimentazione, attendere fino a che il rotore
non si arresta (possono essere necessari alcuni minuti)
Sul lato destro della sede della centrifuga vi è un arresto di plastica (vedere Figura 38)
Rimuovere questo arresto, che è connesso al bloccaggio del coperchio, orizzontalmente dalla
sede fino a che il coperchio della centrifuga non si apre.
Adesso aprire il coperchio della centrifuga
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-21
Figura 38
6.2.2 Descrizione del sistema dei messaggi di errore
Il messaggio di errore "error" (M11) è visualizzato nel display "Time/Prog" (A-3) (Vedere figura 39).
Informazioni dettagliate su possibili messaggi di errore si trovano su "tabella 6: messaggi di errore " (Vedere
Appendice).
Figura 39
7. RICEZIONE DELLE CENTRIFUGHE PER RIPARAZIONE
Rischi per la salute da apparecchiature, rotori e accessori contaminati.
Nel caso di ritorno al costruttore della centrifuga per riparazioni, si prega di notare quanto segue:
La centrifuga deve essere decontaminata e pulita prima della spedizione per la protezione di persone, ambiente e
materiale.
Certificazione di decontaminazione al momento della consegna dei beni da restituire (Vedere APPENDICE)
Ci riserviamo il diritto di non accettare centrifughe contaminate.
Inoltre, tutti i costi sostenuti per la pulizia e la disinfezione dei gruppi saranno addebitati al cliente.
A-3
M11
IT-22 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
8. TRASPORTO, STOCCAGGIO E SMALTIMENTO
8.1 Trasporto
Smontare il rotore prima del trasporto.
Trasportare il gruppo solo nell'imballaggio originale.
Quando si trasporta per lunghe distanze, installare il materiale da
protezione del trasporto per fissare l'albero motore.
Air temperature rel. humidity Air pressure
General transportation -25 to 60 °C 10 to 75 % 30 to 106 kPa
8.2 Stoccaggio
Durante lo stoccaggio della centrifuga, occorre rispettare le condizioni ambientali seguenti:
Air temperature rel. Humidity Air pressure
in transport packaging -25 to 55 °C 10 to 75 % 70 to 106 kPa
8.3 Trasporto, installazione e Smaltimento della Centrifuga
FC5714/FC5718/FC5718R/FC5816/FC5816R
Queste istruzioni complimentano le istruzioni generali precedenti di cui
al capitolo 8 e non le sostituiscono.
8.3.1 Trasporto
Trasportare il gruppo solo nell'imballaggio originale.
La centrifuga deve essere sempre trasportata utilizzando un
dispositivo di trasporto meccanico.
8.3.2 Installazione
Aprire il cartone e sollevare il dispositivo
1. Tagliare il nastro adesivo
2. Aprire le 4 alette del cartone
3. Rimuovere gli accessori
4. Sollevare con cura la centrifuga dal cartone
Avvertenza: Pericolo da sollevamento Vi è rischio di lesioni si sollevata da una sola persona
Utilizzare un dispositivo di sollevamento meccanico oppure procedure di sollevamento in
equipe quando si solleva o si sposta l'apparecchiatura.
Posizionare il dispositivo su un banco da laboratorio stabile, orizzontale e
senza risonanze.
1. Rimuovere il materiale di protezione anteriore e posteriore
2. Rimuovere il manicotto di plastica.
3. Mantenere una distanza minima di 30 cm dai dispositivi in prossimità, ai
lati e dal lato posteriore alla parete.
4. Installare il dispositivo in un luogo ben ventilato, protetto dalla luce solare
diretta, al fine di impedirne il surriscaldamento.
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-23
Connettere il dispositivo.
1. Dopo l'installazione, attendere 4 ore prima di accendere la centrifuga, al
fine di evitare danni al compressore.
2. Verificare che la tensione e la frequenza di rete corrispondano ai requisiti
di targa del dispositivo (Vedere lato posteriore del dispositivo) e quindi
connettere il dispositivo all'alimentazione.
Rimuovere il materiale di protezione durante il trasporto dalla camera del
rotore.
1. Accendere il dispositivo sull'interruttore di alimentazione di rete.
2. Aprire il coperchio della centrifuga utilizzando il pulsante Open
3. Rimuovere il materiale di protezione di trasporto
4. Posizionare il rotore verticalmente sull'albero motore.
5. Ruotare il dado del rotore ruotando la chiave del rotore in senso orario
fino a che il dado del rotore non è serrato.
Adesso il dispositivo é pronto per essere utilizzato.
Conservare l'imballaggio e tutto il materiale di protezione durante il trasporto
per spedire il dispositivo a una data successiva.
8.3.3 Imballaggio
Imballare la centrifuga nell'ordine inverso
8.3.4 Trasferimento del dispositivo
Quando si trasferisce il dispositivo a terzi, accertarsi che comprenda anche il presente manuale di istruzioni.
IT-24 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9. DATI TECNICI
9.1 Specifiche
9.1.1 Centrifuga FC5714
Modello
FC5714
Gamma velocità:
200 giri/minuto -14000 giri/minuto;10 giri/minuto/impostazione
RCF massimo
18624 x g;10 x g/impostazione
Capacità Max. (Rotore)
4x100ml
Gamma Temperature (N/A)
Aria fresca
Tempo di funzionamento
da 10 sec a 99 ore 59 min 59 sec o continuo
Livello di rumorosità (dipendente dal
rotore)
≤ 63 ± 2 dB(A)
Densità consentita a max. velocità
1.2 g/ml
Energia cinetica consentita
5595 Nm
Connessione di alimentazione di rete AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Fluttuazione di tensione
±10 %
±10 %
Consumo di corrente
1,3 A
2,4 A
Consumo di potenza
240 W
300 W
Dimensioni (W × D × H)
362 x 493 x 330 mm
14,3 x 19,4 x 13,0 in
Peso Netto (senza rotore)
30 kg
66 lb
Dimensioni di Spedizione (W × D × H)
580 x 490 x 460 mm
22,8 x 19,3 x 18,1 in
Peso di Spedizione (senza rotore)
32,5 kg
72 lb
Condizioni ambiente (EN/IEC 61010-1)
Ambiente
Solo per uso al coperto
Altitudine
Utilizzabile a un'altitudine fino a 2000 m
Temperatura ambiente
da 2°C a 35 °C
Umidità relativa max.
Umidità relativa max. 80 % per temperature fino a 31°C,
in diminuzione lineare al 50 % di umidità relativa fino a 35°C.
Categoria sovratensione (IEC 60364-4-443)
II
Grado di contaminazione
2
Classe di protezione
I
Non adatta per l'utilizzo in ambienti pericolosi
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-25
9.1.2 Centrifuga FC571
Modello
FC5718
Gamma velocità:
200 giri/minuto -18000 giri/minuto;10 giri/minuto/impostazione
RCF massimo
23542 x g;10 x g/impostazione
Capacità Max. (Rotore)
4x100ml
Gamma Temperature (N/A)
Aria fresca
Tempo di Funzionamento Livello di
rumorosità
da 10 sec a 99 ore 59 min 59 sec o continuo
Livello di rumorosità (dipendente dal
rotore)
≤ 60 ± 2 dB(A)
Densità consentita a max. velocità
1,2 g/ml
Energia cinetica consentita
16672 Nm
Connessione di alimentazione di rete AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Fluttuazione di tensione
±10 %
±10 %
Consumo di corrente
2,0 A
4,0 A
Consumo di potenza
455 W
475 W
Dimensioni (W × D × H)
408 x 499 x 351 mm
16,1 x 19,7 x 13,8 in
Peso Netto (senza rotore)
43 kg
95 lb
Dimensioni di Spedizione (W × D × H)
650 x 520 x 490 mm
25,6 x 20,5 x 19,3 in
Peso di Spedizione (senza rotore)
53 kg
117 lb
Condizioni ambiente (EN/IEC 61010-1)
Ambiente
solo per uso al coperto
Altitudine
Utilizzabile a un'altitudine fino a 2000 m
Temperatura ambiente
da 2°C a 35 °C
Umidità relativa max.
Umidità relativa max. 80 % per temperature fino a 31°C,
in diminuzione lineare al 50 % di umidità relativa fino a 35°C.
Categoria sovratensione (IEC 60364-4-
443)
II
Grado di contaminazione
2
Classe di protezione
I
Non adatta per l'utilizzo in ambienti pericolosi
IT-26 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.3 Centrifuga FC5718R
Modello
FC5718R
Gamma velocità:
200 giri/minuto -18000 giri/minuto;10 giri/minuto/impostazione
RCF massimo
23542 x g;10 x g/impostazione
Capacità Max. (Rotore)
4x100ml
Gamma Temperature (Digitale)
-20~40
Tempo di funzionamento
da 10 sec a 99 ore 59 min 59 sec o continuo
Livello di rumorosità (dipendente dal
rotore)
60 ± 2 dB(A)
Densità consentita a max. velocità
1,2 g/ml
Energia cinetica consentita
25111 Nm
Connessione di alimentazione di rete AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Fluttuazione di tensione
±10 %
±10 %
Consumo di corrente
3,0 A
6,0 A
Consumo di potenza
660 W
660 W
Dimensioni (W × D × H)
407 x 731 x 359 mm
16,0 x 28,8 x 14,1 in
Peso Netto (senza rotore)
60 kg
132 lb
Dimensioni di Spedizione (W × D × H)
840 x 640 x 590 mm
33,1 x 25,2 x 23,2 in
Peso di Spedizione (senza rotore)
77 kg
170 lb
Condizioni ambiente (EN/IEC 61010-1)
Ambiente
solo per uso al coperto
Altitudine
Utilizzabile a un'altitudine fino a 2000 m
Temperatura ambiente
da 2°C a 35 °C
Umidità relativa max.
Umidità relativa max. 80 % per temperature fino a 31°C,
in diminuzione lineare al 50 % di umidità relativa fino a 35°C.
Categoria sovratensione (IEC 60364-4-
443)
II
Grado di contaminazione
2
Classe di protezione
I
Non adatta per l'utilizzo in ambienti pericolosi
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-27
9.1.4 Centrifuga FC5816
Modello
FC5816
Gamma velocità:
200 giri/minuto -15000 giri/minuto;10 giri/minuto/impostazione
RCF massimo
21379 x g;10 x g/impostazione
Capacità Max. (Rotore)
6 x 250ml
Gamma Temperature (N/A)
Aria fresca
Tempo di funzionamento
da 10 sec a 99 ore 59 min 59 sec o continuo
Livello di rumorosità (dipendente dal
rotore)
≤ 61 ± 2 dB(A)
Densità consentita a max. velocità
1,2 g/ml
Energia cinetica consentita
34363 Nm
Connessione di alimentazione di rete AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Fluttuazione di tensione
±10 %
±10 %
Consumo di corrente
2,4 A
4,2 A
Consumo di potenza
530 W
520 W
Dimensioni (W × D × H)
446 x 538 x 354 mm
17,6 x 21,2 x 13,9 in
Peso Netto (senza rotore)
52 kg
115 lb
Dimensioni di Spedizione (W × D × H)
840 x 640 x 590 mm
33,1 x 25,2 x 23,2 in
Peso di Spedizione (senza rotore)
77 kg
170 lb
Condizioni ambiente (EN/IEC 61010-1)
Ambiente
solo per uso al coperto
Altitudine
Utilizzabile a un'altitudine fino a 2000 m
Temperatura ambiente
da 2°C a 35 °C
Umidità relativa max.
Umidità relativa max. 80 % per temperature fino a 31°C,
in diminuzione lineare al 50 % di umidità relativa fino a 35°C.
Categoria sovratensione (IEC 60364-4-
443)
II
Grado di contaminazione
2
Classe di protezione
I
Non adatta per l'utilizzo in ambienti pericolosi
IT-28 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.5 Centrifuga FC5816R
Modello
FC5816R
Gamma velocità:
200 giri/minuto -16000 giri/minuto;10 giri/minuto/impostazione
RCF massimo
24325 x g;10 x g/impostazione
Capacità Max. (Rotore)
6 x 250ml
Gamma Temperature (Digitale)
-20~40
Tempo di funzionamento
da 10 sec a 99 ore 59 min 59 sec o continuo
Livello di rumorosità (dipendente dal
rotore)
≤ 63 ± 2 dB(A)
Densità consentita a max. velocità
1,2 g/ml
Energia cinetica consentita
34363 Nm
Connessione di alimentazione di rete AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Fluttuazione di tensione
±10 %
±10 %
Consumo di corrente
3,7 A
7,8 A
Consumo di potenza
785 W
850 W
Dimensioni (W × D × H)
723 x 538 x 354 mm
28,5 x 21,2 x 13,9 in
Peso Netto (senza rotore)
77 kg
170 lb
Dimensioni di Spedizione (W × D × H)
840 x 640 x 590 mm
33,1 x 25,2 x 23,2 in
Peso di Spedizione (senza rotore)
87 kg
192 lb
Condizioni ambiente (EN/IEC 61010-1)
Ambiente
solo per uso al coperto
Altitudine
Utilizzabile a un'altitudine fino a 2000 m
Temperatura ambiente
da 2°C a 35 °C
Umidità relativa max.
Umidità relativa max. 80 % per temperature fino a 31°C,
in diminuzione lineare al 50 % di umidità relativa fino a 35°C.
Categoria sovratensione (IEC 60364-4-
443)
II
Grado di contaminazione
2
Classe di protezione
I
Non adatta per l'utilizzo in ambienti pericolosi
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-29
9.1.6 Centrifuge FC5916
Modello
FC5916
Intervallo veloci
200 rpm -16000 rpm;10 rpm/impostato
RCF massimo
24325 x g;10 x g/impostato
Capacità massima (rotore)
4X750 ml
Intervallo di temperatura (N/A)
Raffreddamento ad aria
Tempo di esecuzione
10 sec a 99 ore 59 min 59 sec o continuo
Livello di rumore (a seconda del rotore)
63 ± 2 dB(A)
Densità consentita alla massima
velocità
1,2 g/ml
Energia cinetica consentita
60629 Nm
Collegamento alla rete elettrica CA
230 V ~
50/60
Hz
120 V ~ 50/60 Hz
Fluttuazione di tensione
± 10 %
± 10 %
Consumo di corrente
2,8 A
5,6 A
Consumo di energia
640 W
680 W
Dimensioni (L × P × A)
540 x 670 x 390 mm
21,3 x 26,4 x 15,4 pollici
Peso netto (senza rotore)
85 kg
187 libbre
Dimensioni di spedizione (L × P × A)
780 x 670 x 590 mm
30,7 x 26,4 x 23,3 pollici
Peso di spedizione (senza rotore)
98 kg
216 libbre
Condizioni ambientali (EN/IEC 61010-1)
Ambiente
Solo per uso interno
Altitudine
Utilizzare fino ad un'altitudine di 2000 m
Temperatura ambiente
C fino a 35°C
Umidi relativa massima
Umidità relativa max. 80% per temperature fino a 31°C,
diminuendo linearmente fino al 50% di umidità relativa
fino a 35°C.
Categoria di sovratensione (IEC 60364-
4-443)
II
Grado di contaminazione
2
Classe di protezione
I
Non adatto per l'uso in ambienti pericolosi.
IT-30 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
9.1.7 Centrifuge FC5916R
Modello
FC5916R
Intervallo veloci
200 rpm -16000 rpm;10 rpm/impostato
RCF massimo
26331 x g;10 x g/impostato
Capacità massima (rotore)
4X750 ml
Intervallo di temperatura (Digitale)
-20~40
Tempo di esecuzione
10 sec a 99 ore 59 min 59 sec o continuo
Livello di rumore (a seconda del rotore)
63 ± 2 dB(A)
Densità consentita alla massima
velocità
1,2 g/ml
Energia cinetica consentita
54458 Nm
Collegamento alla rete elettrica CA
230 V ~
50/60 Hz
120 V ~ 50/60 Hz
Fluttuazione di tensione
± 10 %
± 10 %
Consumo di corrente
7,2 A
20 A
Consumo di energia
1630 W
1750 W
Dimensioni (L × P × A)
730 x 670 x 390 mm
28,7 x 26,4 x 15,4 pollici
Peso netto (senza rotore)
118 kg
260 libbre
Dimensioni di spedizione (L × P × A)
900 x 750 x 560 mm
40,0 x 29,5 x 22,0 pollici
Peso di spedizione (senza rotore)
137 kg
137 kg
Condizioni ambientali (EN/IEC 61010-1)
Ambiente
solo per uso interno
Altitudine
Utilizzare fino ad un'altitudine di 2000 m
Temperatura ambiente
C fino a 35°C
Umidi relativa massima
Umidità relativa max. 80% per temperature fino a 31°C,
diminuendo linearmente fino al 50% di umidità relativa fino
a 35°C.
Categoria di sovratensione (IEC 60364-
4-443)
II
Grado di contaminazione
2
Classe di protezione
I
Non adatto per l'uso in ambienti pericolosi.
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-31
9.2 Disegni e dimensioni
H
W
W
H
D
D
Modello
L (mm / pollici)
P (mm / pollici)
A (mm / in.)
FC5714
362 / 14.3
493 / 19.4
330 / 13.0
FC5718
408 / 16.1
499 / 19.7
351 / 13.8
FC5718R
408 / 16.1
731 / 28.8
359 / 14.1
FC5816
446 / 17.6
538 / 21.2
354 / 13.9
FC5816R
723 / 28.5
538 / 21.2
354 / 13.9
FC5916
540 / 21.3
670 / 26.4
390 / 15.4
FC5916R
730 / 28.7
670 / 26.4
390 / 15.4
10. CONFORMITA
Il corrispondente marchio di prodotto riporta la conformi alle normative seguenti.
Marcatura Norma
Questo prodotto è conforme con le normative armonizzate applicabili delle
direttive Europee 2011/65/UE (RoHS), 2014/30/UE (EMC) e 2014/35/UE (LVD).
La dichiarazione di Conformi Completa è disponibile on-
line su www.ohaus.com/ce.
Smaltimento
In conformi alla Direttiva Europea 2012/19/UE sui Rifiuti di Apparecchiature
Elettrice ed Elettroniche (RAEE) , questo dispositivo non p essere smaltito
con i
rifiuti domestici . Ciò vale anche per i paesi al di fuori dell'UE, come da requisiti
specifici
Si prega di smaltire il prodotto conformemente ai regolamenti locali al punto di
raccolta specificato per le apparecchiature elettriche ed elettroniche.
Se avete domande, vi invitiamo a prendere contatto con l'autori responsabile
oppure il distributore da cui avete acquistato questo dispositivo.
Se questo dispositivo viene trasmesso ad altre persone (per utilizzo privato o
professionale), anche il contenuto di questo regolamento deve essere trasmesso.
per le istruzioni sullo smaltimento in Europa, fare riferimento a
www.ohaus.com/weee.
Grazie per il vostro contributo alla protezione ambientale.
IT-32 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11. APPENDICE
TABELLA 1: PESO NETTO AMMESSO
TABELLA 2: TEMPERATURE PIÙ BASSE A VELOCITÀ MAX.
TABELLA 3: VELOCITÀ MAX. E VALORI RCF PER ROTORI AMMISSIBILI
TABELLA 4: TEMPI DI ACCELERAZIONE E DECELERAZIONE
TABELLA 5: MESSAGGI DI ERRORE
TABELLA 6; CORREZIONE DEL RAGGIO
TABELLA 7: MODULO DI RISANAMENTO/CERTIFICATO DI DECONTAMINAZIONE
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-33
11.1 Table 1: PESO NETTO AMMESSO
Numero del
rotore
Descrizione del rotore
Peso netto
consentito
Velocità massima ( x1000 rpm)
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
16.0
16.0
30314820
Rotor, Sw ing out, 4x250ml, ID
4x 536 g
4.5
4.5
30314821
Rotor, Angle, 6x250ml FB, ID
6x 355 g
8.0
8.0
8.0
8.0
30314822
Rotor, Sw ing out, 4x100ml, ID
4x 340 g
4.5
4.5
4.5
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
4x 465 g
4.0
5.0
5.0
30314824
Rotor, Sw ing out, 2x3MTP, ID
2x 310 g
4.5
4.5
4.5
4.5
4.5
4.5
4.5
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
6x 140 g
11.0
13.5
30314826
Rotor, Angle, 6x85ml RB, ID
6x 140 g
9.0
9.0
11.0
13.0
11.0
11.0
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
4x 140 g
12.0
12.0
12.0
12.0
15.0
15.0
30314828
Rotor, Sw ing out, 16x50ml, ID
4x 557 g
4.5
4.5
IT-34 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314829
Rotor, Angle, 10x50ml FA, ID
10x 74 g
7.5
7.5
9.0
10.5
10.0
10.0
30314830
Rotor, Angle, 6x50ml RB/FA, ID
6x 72 g
6.0
6.0
6.0
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
6x 94 g
12.0
12.0
13.0
13.0
13.0
13.0
30314832
Rotor, Angle, 30x15ml RB/FA, ID
30x 32 g
4.5
4.5
4.5
4.5
4.5
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
20x 18 g
12.0
12.0
12.0
12.0
12.0
12.0
30314834
Rotor, Angle, 12x15ml RB/FA, ID
12x 25 g
6.0
6.0
6.0
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
44x 3.8 g
13.5
13.5
15.0
16.0
16.0
16.0
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
30x 3.4 g
12.0
13.0
14.0
15.0
15.0
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
24x 3.4 g
14.0
14.0
15.0
15.0
16.0
15.0
15.0
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
12x 3.4 g
18.0
18.0
30314840
Rotor, Angle, 64x 0.5ml, ID
64x 2.3 g
13.5
13.5
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
4x 3.5 g
15.0
15.0
15.0
15.0
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
4x 995 g
4.0
4.0
30553086
Rotor Angle 4x 500ml D69mm ID
4x 708 g
8.0
8.0
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-35
11.2 Table 2: TEMPERATURE PIÙ BASSE A VELOCITÀ MAX.
Numero del rotore
Descrizione del rotore
Modello
Velocità massima
N-massimo
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314821
Rotor, Angle, 6x250ml FB, ID
FC5916R
8000 rpm
-1 ° C
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5718R
4500 rpm
-7 ° C
FC5816R
4500 rpm
-7 ° C
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5718R
5000 rpm
-3 ° C
FC5816R
5000 rpm
-3 ° C
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5718R
4500 rpm
-6 ° C
FC5816R
4500 rpm
-6 ° C
FC5916R
4500 rpm
-8 ° C
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718R
13500 rpm
17 ° C
FC5816R
13500 rpm
17 ° C
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718R
9000 rpm
-5 ° C
FC5816R
9000 rpm
-5 ° C
FC5916R
13000 rpm
8 ° C
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718R
12000 rpm
5 ° C
FC5816R
12000 rpm
5 ° C
FC5916R
16000 rpm
7 ° C
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718R
7500 rpm
9 ° C
FC5816R
7500 rpm
9 ° C
FC5916R
10500 rpm
2 ° C
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5718R
6000 rpm
-8 ° C
FC5816R
6000 rpm
-8 ° C
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718R
12000 rpm
-5 °C
FC5816R
12000 rpm
-5 ° C
FC5916R
13000 rpm
-4 ° C
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5718R
4500 rpm
-9 ° C
FC5816R
4500 rpm
-9 ° C
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718R
12000 rpm
2 ° C
FC5816R
12000 rpm
2 ° C
FC5916R
12000 rpm
-2 ° C
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5718R
6000 rpm
-11.5 ° C
FC5816R
6000 rpm
-11.5 ° C
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718R
13500 rpm
1 ° C
FC5816R
13500 rpm
1 ° C
FC5916R
16000 rpm
3 ° C
30314836
Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable
FC5718R
14000 rpm
9 ° C
FC5816R
14000 rpm
9 ° C
FC5916R
15000 rpm
3 ° C
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5718R
15000 rpm
6 ° C
FC5816R
15000 rpm
6 ° C
FC5916R
16000 rpm
1 ° C
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718R
18000 rpm
-2 ° C
FC5816R
18000 rpm
-2 ° C
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718R
13500 rpm
2 ° C
FC5816R
13500 rpm
2 ° C
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718R
15000 rpm
-1 ° C
FC5816R
15000 rpm
-1 ° C
FC5916R
15000 rpm
-2 ° C
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916R
4500 rpm
4 ° C
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916R
8000 rpm
0 ° C
Tutte le indicazioni di temperatura si riferiscono ad una temperatura ambiente di 23 ° C. Superando questo
valore o la radiazione solare diretta verso la centrifuga, questi valori non possono essere mantenuti.
IT-36 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
Numero del rotore
Descrizione del rotore
Modello
Velocità massim
RPM
RCF
30304361
Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
4500
3780 x g
FC5816R
4500
3780 x g
30314821
Rotor, Angle, 6x250ml FB, ID
FC5816
8000
10016 x g
FC5816R
8000
10016 x g
FC5916
8000
10016 x g
FC5916R
8000
10016 x g
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5714
4500
3350 x g
FC5718
4500
3350 x g
FC5718R
4500
3350 x g
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
4000
2486 x g
FC5718
5000
3885 x g
FC5718R
5000
3885 x g
30314824
Rotor, Sw ing out, 2x3MTP, ID
FC5714
4500
2716 x g
FC5718
4500
2716 x g
FC5718R
4500
2716 x g
FC5816
4500
2716 x g
FC5816R
4500
2716 x g
FC5916
4500
2716 x g
FC5916R
4500
2716 x g
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
11000
13932 x g
FC5718R
13500
20984 x g
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
9000
10413 x g
FC5718R
9000
10413 xg
FC5816
11000
15555 x g
FC5816R
13000
21726 x g
FC5916
11000
15555 x g
FC5916R
13000
21726 x g
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718
12000
14809 x g
FC5718R
12000
14809 x g
FC5816
12000
14809 x g
FC5816R
12000
14809 x g
FC5916
15000
23140 x g
FC5916R
16000
26328 x g
30314828
Rotor, Sw ing out, 16x50ml, ID
FC5816
4500
3735 x g
FC5816R
4500
3735 x g
11.3 Table 3: VELOCITÀ MAX. E VALORI RCF PER ROTORI AMMISSIBILI
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-37
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718
7500
8174 x g
FC5718R
7500
8174 x g
FC5816
9000
11771 x g
FC5816R
10500
16022 xg
FC5916
10000
14532 x g
FC5916R
10500
16022 x g
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
12000
13522 x g
FC5718R
12000
13522 x g
FC5816
13000
15869 x g
FC5816R
13000
15869 x g
FC5916
13000
15869 x g
FC5916R
13000
15869 x g
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5714
4500
2830 x g
FC5718
4500
2830 x g
FC5718R
4500
2830 x g
FC5816
4500
2830 x g
FC5816R
4500
2830 x g
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
12000
15775 x g
FC5718R
12000
15775 x g
FC5816
12000
15775 x g
FC5816R
12000
15775 x g
FC5916
12000
15775 x g
FC5916R
12000
15775 x g
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
6000
4427 x g
FC5718
6000
4427 x g
FC5718R
6000
4427 x g
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718
13500
17113 x g
FC5718R
13500
17113 x g
FC5816
15000
21128 x g
FC5816R
16000
24039 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314836
Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable
FC5714
12000
15131 x g
FC5718
13000
17758 x g
FC5718R
14000
20595 x g
FC5916
15000
23643 x g
FC5916R
15000
23643 x g
IT-38 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5714
14000
18624 x g
FC5718
14000
18624 x g
FC5718R
15000
21379 x g
FC5816
15000
21379 x g
FC5816R
16000
24325 x g
FC5916
16000
24325 x g
FC5916R
16000
24325 x g
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718
18000
23542 x g
FC5718R
18000
23542 x g
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
13500
16298 x g
FC5718R
13500
16298 x g
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
15000
15343 x g
FC5718R
15000
15343 x g
FC5916
15000
15343 x g
FC5916R
15000
15343 x g
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916
4000
3434 x g
FC5916R
4500
4346 x g
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916
8000
10367 x g
FC5916R
8000
10367 x g
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-39
11.4 Table 4: TEMPI DI ACCELERAZIONE E DECELERAZIONE
Articolo n.
Descrizione del rotore
Modello
Accelerazione* Tempo in
sec
Decelerazione* Tempo in
sec
niveau 0
niveau 9
niveau 0
niveau 9
30304361
Rotor, Angle, 24x1.5/2.0ml, ID, Sealable
FC5714
200
25
230
35
FC5718
240
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
20
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314820
Rotor, Sw ing out, 4x250ml, ID
FC5816
309
34
458
36
FC5816R
309
34
458
36
30314821
Rotor, Angle, 6x250ml FB, ID
FC5816
664
130
2906
92
FC5816R
664
130
2906
92
FC5916
573
66
1903
84
FC5916R
573
66
1903
84
30314822
Rotor, Sw ing out, 4x100ml, ID
FC5714
110
14
170
19
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314823
Rotor, Sw ing out, 4x100ml, ID, Sealable
FC5714
110
14
170
17
FC5718
100
15
150
15
FC5718R
100
15
150
15
30314824
Rotor, Sw ing out, 2x 3MTP, ID
FC5714
230
25
340
26
FC5718
160
20
360
15
FC5718R
160
20
360
15
FC5816
452
43
616
38
FC5816R
452
43
616
38
FC5916
249
27
488
23
FC5916R
249
27
488
23
30314825
Rotor, Angle, 6x85ml RB, ID, Hi
FC5718
400
40
960
40
FC5718R
500
60
1260
50
30314826
Rotor, Angle, 6x85ml RB, ID
FC5718
360
40
1050
40
FC5718R
360
40
1050
40
FC5816
697
85
2313
70
FC5816R
2313
70
1630
76
FC5916
463
48
1654
46
FC5916R
549
69
1307
67
30314827
Rotor, Angle, 4x85ml RB, ID, Hi
FC5718
300
60
820
40
FC5718R
300
60
820
40
FC5816
506
60
1745
49
FC5816R
506
60
1745
49
FC5916
448
50
1251
45
FC5916R
448
50
1251
45
30314828
Rotor, Sw ing out, 16x50ml, ID
FC5816
311
34
387
36
FC5816R
311
34
387
36
30314829
Rotor, Angle, 10x50ml FA, ID
FC5718
380
63
796
37
FC5718R
380
62
1069
36
FC5816
753
115
2395
72
FC5816R
753
115
2395
72
FC5916
480
60
1747
68
FC5916R
480
60
1747
68
30314830
Rotor, Angle, 6x50ml RB/FA, ID
FC5714
101
11
206
14
FC5718
89
13
239
11
FC5718R
90
12
463
11
IT-40 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314831
Rotor, Angle, 6x50ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
446
48
1323
49
FC5816R
446
48
1323
49
FC5916
264
28
921
32
FC5916R
264
28
921
32
30314832
Rotor, Angle, 30x15ml RB/FA, ID
FC5714
157
19
370
20
FC5718
160
15
380
10
FC5718R
160
15
380
10
FC5816
149
25
985
20
FC5816R
149
25
985
20
30314833
Rotor, Angle, 20x10ml RB, ID, Hi
FC5718
360
40
570
30
FC5718R
360
40
570
30
FC5816
512
54
1439
47
FC5816R
512
54
1439
47
FC5916
305
32
988
37
FC5916R
305
32
988
37
30314834
Rotor, Angle, 12x15ml RB/FA, ID
FC5714
102
11
167
14
FC5718
88
13
222
11
FC5718R
88
12
433
11
30314835
Rotor, Angle, 44x1.5/2.0ml, ID
FC5718
196
22
204
18
FC5718R
196
22
204
18
FC5816
257
29
436
28
FC5816R
274
31
485
29
FC5916
233
25
283
30
FC5916R
233
25
500
25
30314836
Rotor, Angle, 30x1.5/2.0ml, ID, Sealable
FC5714
250
27
280
34
FC5718
200
25
360
25
FC5718R
210
25
360
30
FC5916
221
23
561
30
FC5916R
221
23
561
30
30314838
Rotor, Angle, 24x1.5/2.0ml, ID
FC5714
200
22
230
35
FC5718
210
23
240
16
FC5718R
230
25
420
17
FC5816
251
25
610
26
FC5816R
249
30
635
27
FC5916
204
21
421
30
FC5916R
204
21
421
30
30314839
Rotor, Angle, 12x1.5/2.0ml, ID
FC5718
240
25
210
20
FC5718R
240
25
210
20
30314840
Rotor, Angle, 64x0.5ml, ID
FC5718
150
17
170
12
FC5718R
150
17
170
12
30314841
Rotor, Angle, 4x8-w PCR Strip, ID
FC5718
130
16
130
12
FC5718R
130
16
130
12
FC5916
100
12
201
12
FC5916R
100
12
201
12
30553085
Rotor Sw ingOut 4x 750ml D99mm ID SL
FC5916
483
47
1287
49
FC5916R
551
55
1501
54
30553086
Rotor Angle 4x 500ml D69mm ID
FC5916
575
73
2317
82
FC5916R
575
73
2317
82
* Nota: accelera da 0 a Vmax; decelera da Vmax a 0.
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-41
11.5 Tableau 5 : MESSAGGI DI ERRORE
Error-No.:
Descri pti on
1
Imbalance arose
2
Imbalance sensor i s defectiv e
4
Imbalance switch has been activated for l onger than 5 secon ds
8
T ransponder i n the rotor i s defectiv e
11
T emperatu re sensor i s defectiv e
12
Chamber over temperature
14
Leap of speed i s too bi g between two mesaureme nts
CLOSE lid
33
Open lid while motor is running
34
Lid contact defectiv e
38
Lid motor is bl ocked
40
Communicatio n wi th frequen cy converter di strubed duri ng start
41
Communicatio n wi th frequen cy converter di strubed duri ng stop
42
Short circuit in the frequency converter
43
Undervoltag e frequen cy converter
44
Overvoltag e frequen cy converter
45
Over temperatu re frequency converter
46
Over temperatu re motor
47
Over current frequen cy conver ter
48
T i meout between control unit and frequency converter
49
Other error frequency converter
55
Overspeed
70
T i meout between controler and RS232 i nterface
99
Rotor i s not allowed i n this centri fuge
FALSE
Inserted rotor does not exi st in the programm
rotor no
Rotor i s not detected
IT-42 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.6 Table 6: CORREZIONE DEL RAGGIO
Rotore
Descrizione del
rotore
Pala/Adatt
atore
Descrizione della pala/adattatore
Raggio
(cm)
Correzione
(cm)
Compatible
FC5714
FC5718
FC5718R
FC5816
FC5816R
FC5916
FC5916R
30304361
Rotor, Angle,
24x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314820
Rotor, Sw ing out,
4x 250ml, ID
Blank
NO Bucket/Adapter
16.7
30314903
Rack, 1x250ml D62mm FB, 2/pk
15.7
-1.0
30314904
Rack, 1x100ml D41mm RB, 2/pk
15.9
-0.8
30314905
Rack, 3x50ml D29mm RB, 2/pk
16.3
-0.4
30314906
Rack, 3x50ml D29mm FA, 2/pk
16.7
0.0
30314907
Rack, 1x50ml D34mm RB, 2/pk
15.8
-0.9
30314908
Rack, 2x50ml D29mm Rim, 2/pk
15.8
-0.9
30314909
Rack, 1x 50ml D29mm Rim, 2/pk
16.1
-0.6
30314910
Rack, 5x25ml D24.5mm RB, 2/pk
16.1
-0.6
30314911
Rack, 3x25ml D24.5mm Rim, 2/pk
15.5
-1.2
30314912
Rack, 9x15ml D17mm RB, 2/pk
16.3
-0.4
30314913
Rack, 7x15ml D17mm FA, 2/pk
16.4
-0.3
30314914
Rack, 7x15ml D17mm Rim, 2/pk
16.1
-0.6
30314915
Rack, 8x10ml D16mm RB, 2/pk
16.3
-0.4
30314916
Rack, 14x5-7ml D13mm RB, 2/pk
15.9
-0.8
30314917
Rack, 10x1.6-7ml D13mm Vac, 2/pk
15.9
-0.8
30304367
Rack, 8x4-10ml D16mm Vac, 2/pk
16.3
-0.4
30304368
Adapter, 9x 1.5/2.0ml D11mm, 2/pk
15.7
-1.0
30314821
Rotor, Angle,
6x 250ml FB, ID
Blank
NO Bucket/Adapter
14.1
30304369
Adapter, 8x 1.5ml D11mm, 2/pk
13.2
-0.9
30304370
Adapter, 5x 10ml D16mm RB, 2/pk
13.3
-0.8
30304371
Adapter, 4x 15ml D17mm FA, 2/pk
13.0
-1.1
30304372
Adapter, 2x 30ml D26mm RB, 2/pk
12.5
-1.6
30304373
Adapter, 1x 50ml D29mm RB, 2/pk
12.0
-2.1
30304374
Adapter, 1x 50ml D30mm FA, 2/pk
11.7
-2.4
30559412
Adapter 15x 1.5/2.0ml D11mm
MTBx 2 V1
30559413
Adapter 1x 250ml D60mm CBCx3 V1
11.8
-0.5
30559414
Adapter 1x 175/225ml D61mm FAx 2
V1
12.8
-2.3
30314822
Rotor, Sw ing out,
4x 100ml, ID
Blank
NO Bucket/Adapter
14.8
30314842
Rack, 1x100ml D46mm RB, 2/pk
14.1
-0.7
30314843
Rack, 1x85ml D38mm RB, 2/pk
14.1
-0.7
30314844
Rack, 1x50ml D34mm RB, 2/pk
14.2
-0.6
30314845
Rack, 1x50ml D29mm RB, 2/pk
14.2
-0.6
30314846
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.7
30314847
Rack, 1x50ml D29mm Rim, 2/pk
14.3
-0.5
30314848
Rack, 1x 30ml D25mm Rim, 2/pk
13.8
-1.0
30314849
Rack, 1x25ml D25mm RB, 2/pk
14.4
-0.4
30314850
Rack, 2x15ml D17mm FA, 2/pk
14.8
0.0
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-43
9.5
13.6
-4.6
-0.5
30314851
Rack, 1x15ml D17mm Rim, 2/pk
14.5
-0.3
30314852
Rack, 7x15ml D17mm RB, 2/pk
14.5
-0.3
30314853
Rack, 9x5-7ml D13mm RB, 2/pk
14.2
-0.6
30314854
Rack, 10x1.5ml D11mm, 2/pk
9.9
-4.9
30314855
Rack, 9x4.5ml D12mm FB, 2/pk
14.0
-0.8
30314856
Rack, 7x5-7ml D13mm RB, 2/pk
11.6
-3.2
30314857
Rack, 7x5-7ml D13mm Vac, 2/pk
14.2
-0.6
30314858
Rack, 5x4-10ml D16mm Sar, 2/pk
14.3
-0.5
30314823
Rotor, Sw ing out,
4x 100ml, ID,
Sealable
Blank
NO Bucket/Adapter
14.6
30314860
Bucket, 100ml w o Cap, Sealable,
2/pk
14.2
-0.4
30314861
Bucket, 100ml w/ Cap, Sealable, 2/pk
14.2
-0.4
30314864
Rack, 1x100ml D40mm RB, 2/pk
13.8
-0.8
30314865
Rack, 1x85ml D38mm RB, 2/pk
14.0
-0.6
30314866
Rack, 1x 50ml D34mm RB, 2/pk
14.0
-0.6
30314867
Rack, 1x50ml D29mm RB, 2/pk
14.0
-0.6
30314868
Rack, 1x50ml D29mm FA, 2/pk
14.1
-0.5
30314869
Rack, 1x50ml D29mm Rim, 2/pk
13.9
-0.7
30314870
Rack, 1x30ml D25mm Rim, 2/pk
13.1
-1.5
30314871
Rack, 1x25ml D25mm RB, 2/pk
14.0
-0.6
30314872
Rack, 1x15ml D17mm FA, 2/pk
14.1
-0.5
30314873
Rack, 2x15ml D17mm FA, 2/pk
14.1
-0.5
30314874
Rack, 2x15ml D17mm Rim, 2/pk
14.0
-0.6
30314875
Rack, 4x15ml D17mm RB, 2/pk
14.0
-0.6
30314876
Rack, 7x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314877
Rack, 5x1.5ml D11mm, 2/pk
14.0
-0.6
30314878
Rack, 3x10ml D17mm Sar, 2/pk
14.0
-0.6
30314879
Rack, 4x5-7ml D13mm RB, 2/pk
14.0
-0.6
30314880
Rack, 4x10ml D17mm Vac, 2/pk
14.0
-0.6
30314881
Bucket, 2x50ml D29mm FA, w / Rack,
2/pk
14.6
0.0
30314882
Bucket, 3x15ml D17mm FA, w / Rack,
2/pk
14.6
0.0
30314883
Cy to, Rectangular Bucket, 2/pk
6.3
-8.3
30314824
Rotor, Sw ing out,
2x 3MTP, ID
Blank
NO Bucket/Adapter
12.0
30314891
Adapter, 48x 1.5/2.0ml D11mm, 2/pk
12.0
0.0
30314825
Rotor, Angle,
6x 85ml RB, ID, Hi
Blank
NO Bucket/Adapter
10.3
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
9.6
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
9.6
-0.7
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.0
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
9.8
-0.5
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
9.3
-1.0
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
10.3
0.0
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
9.5
-0.8
30314826
Rotor, Angle,
6x 85ml RB, ID
Blank
NO Bucket/Adapter
11.5
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.9
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
10.4
-1.1
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
10.9
-0.6
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
10.6
-0.9
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
10.4
-1.1
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
11.1
-0.4
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
10.4
-1.1
IT-44 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
30314827
Rotor, Angle,
4x 85ml RB, ID, Hi
Blank
NO Bucket/Adapter
9.2
30314893
Adapter, 1x 15ml D17mm RB, 2/pk
8.5
-0.7
30314894
Adapter, 1x 30ml D25mm RB, 2/pk
8.4
-0.8
30314895
Adapter, 1x 50ml D29mm RB, 2/pk
8.9
-0.3
30314896
Adapter, 1x 50ml D29mm FA, 2/pk
8.6
-0.6
30314897
Adapter, 1x 15ml D17mm FA, 2/pk
8.3
-0.9
30314898
Adapter, 2x 15ml D17mm RB, 2/pk
7.5
-1.7
30314899
Adapter, 1x 16ml D18mm RB, 2/pk
8.3
-0.9
30314828
Rotor, Sw ing out,
16x 50ml, ID
Blank
NO Bucket/Adapter
16.5
30304375
Rack, 4x50ml D29mm FA, 2/pk
16.5
0.0
30314583
Rack, 10x15ml D17mm FA, 2/pk
16.5
0.0
30314829
Rotor, Angle,
10x 50ml FA, ID
Blank
NO Bucket/Adapter
13.0
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
12.8
-0.2
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.4
-2.6
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
8.9
-4.1
30314830
Rotor, Angle,
6x 50ml RB/FA, ID
Blank
NO Bucket/Adapter
11.0
30130891
Adapter, 1x 30ml D26mm, 2/pk
10.7
-0.3
30130892
Adapter, 1x 16ml D18mm, 2/pk
10.3
-0.7
30130893
Adapter, 1x 15ml D17mm, 2/pk
10.6
-0.4
30130894
Adapter, 1x 15ml D17mm RB, 2/pk
10.6
-0.4
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.2
-0.8
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
8.3
-2.7
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
6.7
-4.3
30314831
Rotor, Angle,
6x 50ml RB, ID, Hi
Blank
NO Bucket/Adapter
8.4
30130891
Adapter, 1x 30ml D26mm, 2/pk
8.2
-0.2
30130892
Adapter, 1x 16ml D18mm, 2/pk
7.9
-0.5
30130893
Adapter, 1x 15ml D17mm, 2/pk
8.0
-0.4
30314892
Adapter, 1x 15ml D17.5mm FA, 2/pk
7.7
-0.7
30314832
Rotor, Angle,
30x 15ml RB/FA,
ID
Blank
NO Bucket/Adapter
12.5
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
12.2
-0.3
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
10.5
-2.0
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
9.0
-3.5
30314834
Rotor, Angle,
12x 15ml RB/FA,
ID
Blank
NO Bucket/Adapter
11.0
30130889
Adapter, 1x 7ml D13.5mm RB, 2/pk
10.6
-0.4
30130890
Adapter, 1x 5ml D13.5mm, 2/pk
9.1
-1.9
30130886
Adapter, 1x 1.5/2.0ml D11mm, 6/pk
7.7
-3.4
30314835
Rotor, Angle,
44x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.7
-0.7
30314836
Rotor, Angle,
30x 1.5/2.0ml, ID,
Sealable
Blank
NO Bucket/Adapter
9.4
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
9.1
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
8.4
-1.0
30314838
Rotor, Angle,
24x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
8.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
8.2
-0.3
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
7.5
-1.0
30314839
Rotor, Angle,
12x 1.5/2.0ml, ID
Blank
NO Bucket/Adapter
6.5
30130884
Adapter, 1x 0.2/0.4ml D6mm, 6/pk
6.3
-0.2
30130885
Adapter, 1x 0.5ml D8mm, 6/pk
5.6
-0.9
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-45
30314900
Adapter, 1x 1.5ml for ov er16000x g,
6/pk
6.4
-0.1
30553085
Rotor Sw ingOut
4x 750ml D99mm
ID SL
Blank
NO Bucket/Adapter
19.3
30553138
Adapter 32x 7ml D13mm RBx 2 V1
18.7
-0.6
30553135
Adapter 18x 15ml D17mm RBx 2 V1
18.8
-0.5
30553132
Adapter 8x 30ml D26mm RBx 2 V1
19.1
-0.2
30553131
Adapter 7x 50ml D29mm RBx 2 V1
19.1
-0.2
30553127
Adapter 3x 100ml D41mm RBx 2 V1
19.1
-0.2
30553125
Adapter 1x 250ml D62mm FBx 2 V1
18.9
-0.4
30553126
Adapter 1x 225/190ml D61mm FAx 2
V1
19.1
-0.2
30553128
Adapter 5x 50ml D29mm FAx 2 V1
19.1
-0.2
30553130
Adapter 7x 50ml D29mm FAx 2 V1
19.1
-0.2
30553133
Adapter 14x 15ml D17mm FAx 2 V1
19.2
-0.1
30553134
Adapter 17x 15ml D17mm FAx 2 V1
19.0
-0.3
30553140
Adapter 18x 5/10ml D16mm
Vac/Sarx 2 V1
19.0
-0.3
30559377
Adapter 38x 1.1/1.4ml D8mm Sarx2
V1
18.9
-0.4
30553136
Adapter 21x 10ml D16mm RBx 2 V1
18.9
-0.4
30553139
Adapter 27x 1.8/7ml D13mm
Vac/Sarx 2 V1
18.8
-0.5
30553086
Rotor Angle
4x 500ml D69mm
ID
Blank
NO Bucket/Adapter
14.5
30564850
Adapter 1x 250ml D61.5mm FBx 2 V1
13.7
-0.8
30559417
Adapter 1x 175/225ml D61.5mm
FAx 2 V1
13.4
-1.1
30559419
Adapter 1x 50ml D29.5mm FAx 2 V1
12.4
-2.1
30559416
Adapter 1x 250ml D60.5mm CBCx 2
V1
12.6
-1.9
30559421
Adapter 4x 30ml D25.5mm RBx 2 V1
13.8
-0.7
30559420
Adapter 3x 50ml D29mm RBx 2 V1
14.3
-0.2
30559422
Adapter 6x 15ml D17mm FAx 2 V1
13.8
-0.7
IT-46 Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
11.7 Tabella 7: Modulo di risanamento/Certificato di decontaminazione
Allegare il presente modulo a tutti i ritorni di apparecchiature e gruppi
La dichiarazione completa della decontaminazione è un requisito per la ricezione e l'ulteriore elaborazione del ritorno. Se non si allega la
spiegazione corrispondente, eseguiremo la decontaminazione a vostre spese.
Cognome ___________________________________________
Organizzazione/Società ________________________________________
Via: ______________________________________________________
C.A.P. _____________ luogo ___________________
Tel.: _____________ Fax:____________________
E-mail: _____________________________________
Pos. Crowd
Decontaminated
object
Serial number
Description /
Comment
1
2
3
4
Gli elementi elencati qui sopra sono in contatto con le sostanze seguenti?
Soluzioni acquose dannose per la salute, soluzioni tampone, acidi, alcali.... No
Agenti potenzialmente infettivi …………………………………………………………… Sì No
Reagenti organici e solvente ……………………………………………………….. .. No
Sostanze radioattive: ……………………………………………. α.. β.. γ No
Proteine dannose per la salute ………………………………………………………….. No
DNA: ……………………………………………………………………………………… No
Queste sostanze hanno raggiunto l'apparecchiatura/il gruppo ? ………………… No
Se sì, quali :
________________________________________________________________
Descrizione delle misure per la decontaminazione delle parti elencate:
Confermo l'appropriata decontaminazione
Società/Servizio.____________ Luogo e Data: ____________________________________
Firma della persona autorizzata:________________________________________
Si prega di compilare
in stampatello
Frontier
TM
Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-47
Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054 USA
With offices worldwide / Con oficinas en todo el mundo/ Avec des bureaux dans le monde entier/ Mit
Niederlassungen weltweit/ Con uffici in tutto il mondo
http://www.ohaus.com
*30323917*
P/N 30323917 C © 2020 OHAUS Corporation, all rights reserved / todos los derechos reservados / tous droits
réservés / Alle Rechte vorbehalten / Tutti i diritti riservati
Printed in China / Impreso en China/ Imprimé en Chine / Gedruckt in China / Stampato in Cina

Transcripción de documentos

FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Instruction Manual Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Manual de instrucciones Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Manuel d’instructions FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Bedienungsanleitung FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Manuale di Istruzioni Sketch map of FC5714 / FC5718 / FC5816 / FC5916 1 2 7 4 3 6 5 Figure.1 Sketch map of FC5718R 1 2 7 3 4 6 5 Figure.2 Sketch map of FC5816R / FC5916R 1 2 7 3 6 4 5 Figure.3 1 Centrifuge Lid 3 Display 5 Main Power Switch 7 Emergency Release 2 Rotor Window 4 Function Label 6 Power Connection Function Label for FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figure.4 1 4 5 6 7 8 2 3 9 10 11 12 Figure.5 1. 3. 5. 7. 9. LCD Display Start centrifugation Acceleration/Deceleration intensity model and select Time setup model Short/quick spin centrifugation 11. Stop centrifugation / setup 2. 4. 6. Program setup model RPM/RCF model and select Temperature setup model (Only FC5515R) 8. Store setup information 10. Release lid 12. Adjusting knob/Dial: Change the number LCD Display The following picture shows the individual elements of the LCD-display. Figure.6 M2 M3 M1 M12 M15 M13 M4 M9 M11 M10 M5 M6 M7 A-1 M8 A-2 M14 A-3 Figure.7 Display fields: A-1 A-2 A-3 A-4 Display field – "RPM/RCF" Display field – "Acc/Dec" "Service" Display field – "Time/Prog" Display field –"Temp" Messages/logos of the display fields M1 M4 M7 M10 M13 "close" "Rotor-No." "accel" "program" "h m s" M2 M5 M8 M11 M14 "open" "rpm" "decel" "error" "temperature" M3 M6 M9 M12 M15 "rotor" "rcf" "radius" "service" "precool" A-4 Rotor Information Table FC5916 FC5916R Rotor, Angle, 24x1.5/2.0ml, ID, Sealable Rotor, Swing out, 4x250ml, ID Rotor, Angle, 6x250ml FB, ID Rotor, Swing out, 4x100ml, ID Rotor, Swing out, 4x100ml, ID, Sealable Rotor, Swing out, 2x3MTP, ID Rotor, Angle, 6x85ml RB, ID, Hi Rotor, Angle, 6x85ml RB, ID Rotor, Angle, 4x85ml RB, ID, Hi Rotor, Swing out, 16x50ml, ID Rotor, Angle, 10x50ml FA, ID Rotor, Angle, 6x50ml RB/FA, ID Rotor, Angle, 6x50ml RB, ID, Hi Rotor, Angle, 30x15ml RB/FA, ID Rotor, Angle, 20x10ml RB, ID, Hi Rotor, Angle, 12x15ml RB/FA, ID Rotor, Angle, 44x1.5/2.0ml, ID Rotor, Angle, 30x1.5/2.0ml, ID, Sealable Rotor, Angle, 24x1.5/2.0ml, ID Rotor, Angle, 12x1.5/2.0ml, ID Rotor, Angle, 64x0.5ml, ID Rotor, Angle, 4x8-w PCR Strip, ID Rotor SwingOut 4x750ml D99mm ID SL Rotor Angle 4x500ml D69mm ID FC5816R 30304361 30314820 30314821 30314822 30314823 30314824 30314825 30314826 30314827 30314828 30314829 30314830 30314831 30314832 30314833 30314834 30314835 30314836 30314838 30314839 30314840 30314841 30553085 30553086 FC5816 61 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 38 39 40 41 85 86 Capacity FC5718R Order No. FC5718 Rotor No. display FC5714 Compatible ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● TABLE OF CONTENTS 1. 2. 3. 4. 5. INTRODUCTION ............................................................................................................................. 1 1.1 DESCRIPTION AND INTENDED PURPOSE...................................................................................... 1 1.2 BRIEF DESCRIPTION .................................................................................................................... 1 1.3 DEFINITION OF SIGNAL WARNINGS AND SYMBOLS....................................................................... 1 1.4 SAFETY PRECAUTIONS...................................................................................................................................................... 2 1.4.1 USER................................................................................................................................................................................. 2 1.4.2 ROTOR AND ACCESSORIES .................................................................................................................................... 2 1.4.3 MEASURES FOR YOUR PROTECTION.................................................................................................................. 2 1.4.4 EXCLUDE THE FOLLOWING ENVIRONMENTAL INFLUENCES..................................................................... 2 1.4.5 MEASURES FOR OPERATIONAL SAFETY ........................................................................................................... 3 1.4.6 DANGER AND PRECAUTIONS.................................................................................................................................. 3 1.4.7 ABBREVIATIONS USED IN THIS MANUAL............................................................................................................ 3 INSTALLATION ......................................................................................................................................................................... 4 2.1 UNPACKING ............................................................................................................................................................................ 4 2.1.1 DELIVERY PACKAGE................................................................................................................................................... 4 2.2 SELECTING THE LOCATION ............................................................................................................................................. 4 2.3 INSTALLATION ....................................................................................................................................................................... 4 2.4 SAFETY PRECAUTIONS DURING OPERATION AND WARRANTY ....................................................................... 5 OPERATION .................................................................................................................................. 5 3.1 MOUNTING AND LOADING ROTOR ................................................................................................................................ 5 3.1.1 INSTALLATION OF ROTORS ..................................................................................................................................... 5 3.1.2 LOADING ANGLE ROTORS........................................................................................................................................ 6 3.1.3 LOADING SWING OUT ROTORS.............................................................................................................................. 6 3.1.4 LOADING AND OVERLOADING OF ROTORS ...................................................................................................... 7 3.1.5 REMOVING THE ROTOR ............................................................................................................................................ 7 3.2 LID CONTROL......................................................................................................................................................................... 7 3.2.1 LID OPEN......................................................................................................................................................................... 7 3.2.2 LID LOCK.......................................................................................................................................................................... 8 3.3 PRESELECTION..................................................................................................................................................................... 8 3.3.1 PRESELECTION OF SPEED / RCF-VALUE ........................................................................................................... 8 3.3.2 PRESELECTION OF RUNNING TIME...................................................................................................................... 9 3.3.3 PRESELECTION OF BRAKE INTENSITY AND ACCELERATION.................................................................... 9 3.3.4 PRE-SELECTION OF TEMPERATURE (ONLY REFRIGERATED MODELS).............................................. 10 3.3.5 PRE-COOLING (ONLY REFRIGERATED MODELS).......................................................................................... 10 3.4 RADIUS CORRECTION ...................................................................................................................................................... 10 3.5 PROGRAM.............................................................................................................................................................................. 11 3.5.1 STORAGE OF PROGRAMS...................................................................................................................................... 11 3.5.2 RECALL OF STORED PROGRAMS ....................................................................................................................... 12 3.5.3 LEAVING PROGRAM MODE .................................................................................................................................... 12 3.6 STARTING AND STOPPING THE CENTRIFUGE........................................................................................................ 12 3.6.1 STARTING THE CENTRIFUGE................................................................................................................................ 12 3.6.2 THE "STOP" KEY ......................................................................................................................................................... 13 3.7 IMBALANCE DETECTION.................................................................................................................................................. 13 SETTING ................................................................................................................................................................................... 14 4.1 BASIC ADJUSTMENTS....................................................................................................................................................... 14 4.1.1 ACCESS TO MODE "OPERATING DATA" ............................................................................................................ 14 4.1.2 TEMPERATURE INDICATION .................................................................................................................................. 14 4.1.3 SIGNAL TURN ON / OFF ........................................................................................................................................... 15 4.1.4 VOLUME PRE-SELECTION OF SOUND SIGNAL............................................................................................... 15 4.1.5 SONG SELECTION FOR SOUND SIGNAL - END OF RUN ............................................................................. 16 4.1.6 KEYBOARD SOUND TURN ON / OFF ................................................................................................................... 16 4.1.7 CALL UP OPERATING DATA ................................................................................................................................... 16 MAINTENANCE....................................................................................................................................................................... 17 5.1 MAINTENANCE AND CLEANING .................................................................................................................................... 17 5.1.1 GENERAL...................................................................................................................................................................... 17 5.1.2 CLEANING AND DISINFECTION OF THE UNIT ................................................................................................. 18 5.1.3 CLEANING AND DISINFECTION OF THE ROTOR ............................................................................................ 18 5.1.4 DISINFECTION OF ALUMINUM ROTORS ............................................................................................................ 18 5.1.5 DISINFECTION OF PP-ROTORS ............................................................................................................................ 18 5.1.6 GLASS BREAKAGE..................................................................................................................................................... 19 5.2 LIFE TIME OF ROTORS, BUCKETS, ACCESSORIES............................................................................................... 19 TROUBLESHOOTING ........................................................................................................................................................... 19 6.1 ERROR MESSAGE: CAUSE / SOLUTION..................................................................................................................... 19 6.2 SURVEY OF POSSIBLE ERROR MESSAGES AND THEIR SOLUTIONS............................................................ 19 6.2.1 LID RELEASE DURING POWER FAILURE (EMERGENCY LID RELEASE)................................................ 19 6.2.2 DESCRIPTION OF THE ERROR MESSAGE SYSTEM ..................................................................................... 20 7. RECEIPT OF CENTRIFUGES TO REPAIR ..................................................................................................................... 21 8. TRANSPORT, STORAGE AND DISPOSAL..................................................................................................................... 21 8.1 TRANSPORT ......................................................................................................................................................................... 21 8.2 STORAGE............................................................................................................................................................................... 21 8.3 TRANSPORTING, INSTALLING, TRANSFERRING AND DISPOSING OF THE CENTRIFUGE FC5714/FC5718/ FC5718R/FC5816/FC5816R.................................................................................................................... 21 8.3.1 TRANSPORT................................................................................................................................................................. 21 8.3.2 INSTALLATION ............................................................................................................................................................. 21 8.3.3 PACKING........................................................................................................................................................................ 22 8.3.4 PASSING ON THE DEVICE ...................................................................................................................................... 22 9. TECHNICAL DATA ................................................................................................................................................................. 23 9.1 SPECIFICATIONS ................................................................................................................................................................ 23 9.1.1 CENTRIFUGE FC5714................................................................................................................................................ 23 9.1.2 CENTRIFUGE FC5718................................................................................................................................................ 24 9.1.3 CENTRIFUGE FC5718R............................................................................................................................................. 25 9.1.4 CENTRIFUGE FC5816................................................................................................................................................ 26 9.1.5 CENTRIFUGE FC5816R............................................................................................................................................. 27 9.2 DRAWINGS AND DIMENSIONS....................................................................................................................................... 30 10. COMPLIANCE ......................................................................................................................................................................... 31 11. APPENDIX ................................................................................................................................................................................ 33 11.1 TABLE 1: PERMISSIBLE NET WEIGHT....................................................................................................................... 34 11.2 TABLE 2: LOWEST TEMPERATURES AT MAX. SPEED (ONLY REFRIGERATED MODELS).................... 36 11.3 TABLE 3: MAX. SPEED AND RCF-VALUES FOR PERMISSIBLE ROTORS ..................................................... 37 11.4 TABLE 4: ACCELERATION AND DECELERATION TIMES .................................................................................... 40 11.5 TABLE 5: ERROR MESSAGES ...................................................................................................................................... 42 11.6 TABLE 6: RADIUS CORRECTION AND ADAPTER SPECIFICATIONS .............................................................. 43 11.7 TABLE 7: REDEMPTION FORM / DECONTAMINATION CERTIFICATE ............................................................ 47 6. FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-1 1. INTRODUCTION 1.1 Description and Intended Purpose Thank you for choosing this OHAUS product. All symbols indicate safety instructions and points to potential dangerous situations. Please read the manual completely before using the FrontierTM MultiPro Centrifuges to avoid incorrect operation. FrontierTM MultiPro Centrifuges were designed for the separation of materials or mixtures with different densities. OHAUS centrifuges are intended exclusively for indoor use and for use by qualified personnel. 1.2 Brief description The models FC5714/FC5718/FC5816/FC5916 are non-refrigerated universal centrifuges. The models FC5718R/FC5816R/FC5916R are refrigerated universal centrifuges. All models are offered in two voltage variations, 230V or 120V. The centrifuges can be used with swing-out and angle rotors. All parameters are accessible via buttons and selected with the central adjuster. All pre-selected and current values will be shown permanently on the LCD-display. The centrifuge is powered by a maintenance-free induction motor. Detailed technical data are in the "Technical data" section. 1.3 Definition of Signal Warnings and Symbols Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results. The degree of danger is a part of a safety note and distinguishes the possible results of non-observance from each other. Signal Words DANGER WARNING CAUTION ATTENTION NOTE Will lead to severe injuries or death if not avoided. For a hazardous situation with medium risk, possibly resulting in injuries or death if not avoided. For a hazardous situation with low risk, resulting in damage to the device or the property or in loss of data, or injuries if not avoided. For important information about the product. May lead to equipment damage if not avoided For useful information about the product Warning Symbols General Hazard Electrical Shock Hazard Alternating Current Biohazard Explosion Crushing EN-2 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Warning and information signs on the surface of centrifuge Four carriers must be used at all times on four place swing out rotors or damage will occur to the centrifuge. Such damage will not be covered under the product warranty. Attention! Check the fastening of the rotor nut before each run. Take off mains plug before opening the housing or the emergency release. Direction of rotation – clockwise rotation for the rotor drive Reference for loading rotors 1.4 Safety Precautions 1.4.1 User OHAUS centrifuges are intended exclusively for indoor use and for use by qualified personnel. This device may only be operated by trained specialist staff. They must have carefully read the operating manual and be familiar with the functions of the device. 1.4.2 Rotor and accessories Only OHAUS original rotors and accessories shall be used. Any other use or intended use is considered improper. OHAUS is not liable for damage resulting from improper use. CAUTION: Read all safety warnings before installing, making connections, or servicing this equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain instructions for future reference. 1.4.3 Measures for your protection WARNING: Never work in an environment subject to explosion hazards! The housing of the instrument is not gas tight. (Explosion hazard due to spark formation, corrosion caused by the ingress of gases) WARNING: When using chemicals and solvents, comply with the instructions of the producer and the general lab safety rules. WARNING: The centrifuge is not sealed. Use suitable protection measures when using the centrifuge for infectious and pathogenic samples. Follow appropriate safety precautions when handling these samples. 1.4.4 Exclude the following environmental influences  Powerful vibrations  Direct sunlight  Atmospheric humidity greater than 80%  Corrosive gases present  Temperatures below 2 °C and above 35 °C  Powerful electric or magnetic fields: FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-3 WARNING: Electrical shock hazards exist within the housing. The housing should only be opened by authorized and qualified personnel. Remove all power connections to the unit before opening.              1.4.5 Measures for operational safety Do not unscrew the two halves of the housing Dry off any liquid spills immediately! The instrument is not watertight Verify that the equipment’s input voltage range and plug type are compatible with the local power supply. Only connect the power cord to a properly grounded power receptacle. Only use a power cord with a rating that exceeds the specifications on the equipment label. Do not position the equipment such that it is difficult to disconnect the power cord from the power receptacle. Make sure that the power cord does not pose a potential obstacle or tripping hazard. The equipment is for indoor use only. Use the equipment only in dry locations. Use only approved accessories. Operate the equipment only under ambient conditions specified in these instructions. Disconnect the equipment from the power supply when cleaning. Do not operate the equipment in hazardous or unstable environments. Service should only be performed by authorized personnel. 1.4.6 Danger and precautions To protect people and environment the following precautions should be observed:  During centrifugation, the presences of people and the setting up of hazardous materials are prohibited within 30 cm around the centrifuge according to the regulations of EN 61010-2-020.  FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R are not explosion-proof and must therefore not be operated in explosion-endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. The final decision on the risks associated with the use of such substances is the responsibility of the user of the centrifuge.  Never spin toxic or pathogenic material without adequate safety precautions, i.e. centrifugation of buckets / tubes with missing or defective hermetic sealing is strictly prohibited. The user is obliged to perform appropriate disinfection procedures in case dangerous substances have contaminated the centrifuge and or its accessories. When centrifuging infectious substances, always pay attention to the general laboratory precautions. If necessary, contact your safety officer!  It is prohibited to run the centrifuge with rotors other than listed for this unit.  Under no circumstances open the lid of the centrifuge while the rotor is still running or rotating with a speed of > 2m/s 1.4.7 Abbreviations used in this manual Symbol/Abbreviations Unit [min-1] RPM rpm RCF [x g] Description PCR Polymerase chain reaction revolutions per minute relative centrifugal force PP - Polypropylene PC accel decel - Polycarbonate acceleration deceleration prog - program EN-4 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 2. INSTALLATION 2.1 Unpacking Carefully remove your centrifuge and each of its components from the package. The included components vary depending on the centrifuge model (see table below). Save the packaging to ensure safe storage and transport. The instruction manual must always be kept with the centrifuge! Rotor(s) / Accessories will be packed separately. WARNING: Lifting Hazard. Single person lift could cause injury. Use a mechanical lifting device or team lifting procedures when lifting or moving the equipment. Please refer to section 8.3 for details about lifting the equipment out of the packaging. 2.1.1 Delivery package Quantity 1 1 1 1 1 Description Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Power Cable Warranty Card Instruction Manual/Quick Guide Rotor Key 2.2 Selecting the Location NOTE! Avoid excessive vibrations, heat sources, air current, or rapid temperature changes.    The centrifuge should be installed on an even, solid and level surface, if possible on a laboratory cabinet / table or some other solid vibration free surface. During centrifugation, the centrifuge must be placed in a way, that there is a minimum space of 30 cm on each side of the unit according to the standards EN 61010-2-020. Do not place the centrifuge next to a window or a heater, where it could be exposed to excessive heat, as the performance of the unit is based on an ambient temperature of 23°C. 2.3 Installation Follow these steps:  Check whether the power supply corresponds with the one specified on the manufacturer's rating label, which is located on the rear panel.  For FC5714/FC5718/FC5816, the power line should be protected by a 10 A rating circuit breaker (type K).  For FC5718R/FC5816R /FC5916/FC5916R, the power line should be protected by a 16 A rating circuit breaker (type K).  In case of emergency, there must be an emergency switch off installed outside the room in order to disconnect the power supply from the unit.  Connect the centrifuge to a grounded power receptacle.  Connect the centrifuge with the mains. (The socket for the power cord must be easy to reach for disconnection)  Turn the instrument on using the mains power switch.  Open the lid by using the Door Open button.  Remove the transport securing device of the motor. FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-5 2.4 Safety precautions during operation and warranty           Do not operate the centrifuge in case it is not installed correctly. Do not lean on the centrifuge during operation. Do not stay within the 30 cm clearance envelope longer than necessary for operational reasons. Do not place any potentially hazardous materials within the 30 cm clearance envelope. Do not operate the centrifuge when disassembled (e.g. without housing). Do not run the centrifuge when mechanical or electrical components have been tampered with. Do not use accessories such as rotors and buckets, which are not exclusively approved by OHAUS Corporation, except commercially available centrifuge tubes made of glass or plastic. Do not spin extremely corrosive substances, as they may damage or weaken the materials. Do not operate the centrifuge with rotors or buckets, which show any signs of corrosion or mechanical damage. The manufacturer is responsible for safety and reliability of the centrifuge, only if: 1) The unit is operated in accordance with this instruction manual. 2) Modifications, repairs or other adjustments are performed by OHAUS authorized personnel and the electrical installation complies with the relevant electrical code. NOTE! Warranty The centrifuge has been subjected to thorough testing and quality controls. In the unlikely case of any manufacturing faults occurring, the centrifuge and rotors are covered by warranty. This warranty becomes invalid in case of mishandling, damage and negligence and further in case of usage of inappropriate spare parts and / or accessories or unauthorized modification of the unit. Technical modification rights are reserved by the manufacturer in respect to technical improvement! 3. OPERATION 3.1 Mounting and loading rotor 3.1.1 Installation of rotors Clean the drive shaft as well as the collet with a clean, grease-free piece of cloth. Place the rotor onto the drive shaft. (See figure below). Take care that the rotor is fully installed onto the motor shaft. Motor shaft and chamber Figure.8 Nut for Rotor Tool for rotor with nut Figure. 9 Tool for rotor without nut Snap-on lid Screw-on lid Figure.10 EN-6 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Hold the rotor with one hand and secure the rotor to the shaft by turning the fixing nut clockwise. Tighten the fixing nut with enclosed rotor key (See figures 9-10) We will provide a tool for none-nut rotor with centrifuge, the tool for nut-rotor will be provided with rotor. ATTENTION! Check that the fixing screw is properly installed before each run. (See figure 9-10) Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or mechanical damage. Do not operate with extremely corrosive substances, which could damage the rotor, buckets and materials. In case of any questions, please contact the manufacturer! 3.1.2 Loading angle rotors Rotors must be loaded symmetrically and with equal weight (See figure below). The adapter may only be loaded with the appropriate vessels. The weight differences between the filled vessels should be kept as low as possible. Therefore we recommend weighing them with a balance. This reduces the wear of the drive and the acoustic operating noise. The maximum load per hole is stated on each rotor. Figure.11 WRONG Figure.12 CORRECT ( 6 tubes) 3.1.3 Loading swing out rotors Loading of the buckets / vessels must be made in accordance to the figure below. It is allowed to operate e.g. a 4-place-rotor with 2 loaded buckets only. But the loaded buckets must be opposite to each other. Make sure that the unloaded buckets also be put inside the rotor (see below). In principle swing out rotors may not be taken into operation until all buckets or racks are put into the rotor. The bolts at the rotor must be greased with the "High TEF oil". The sample tubes have to be filled evenly by eye and put into the drillings or tube racks. The weight difference of the loaded buckets should not exceed approx.1.0 g. ATTENTION! Swing out rotors may be taken in operation only if all locations are filled in with either four buckets or four carriers – do not mix buckets and carriers together!! ATTENTION! Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or mechanical damage. Do not operate with extremely corrosive substances, which could damage the rotor and buckets. In case of any questions, please contact the manufacturer! FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figure.13 WRONG EN-7 Figure.14 CORRECT 3.1.4 Loading and overloading of rotors All approved rotors are listed with their maximum speed and maximum filling weight in "table 2 permissible net weight" (See APPENDIX). The maximum load permitted for a rotor, which is determined by the manufacturer, as well as the maximum speed allowed for this rotor (See label on rotor), must not be exceeded. The liquids the rotors are loaded with should have a maximum homogeneous density of 1.2 g/ml or less when the rotor is running at maximum speed. In order to spin liquids with a higher density, the speed has to be reduced according to the following formula: Reduced speed nred = √ 1,2 ℎ𝑖𝑔ℎ𝑒𝑟 𝑑𝑒𝑛𝑠𝑖𝑡𝑦 x max. speed (nmax) of the rotor Example: nred = √ 1,2 1,7 x 4.000 = 3.360 rpm If in case of any questions, please contact the manufacturer! To determine the relative centrifugal force(RCF/G-force) for a specific adapter, you can calculate per DIN 58 970 using the attached formula: RCF = 1.117862*10-5*n2*rmax n: revolutions per minute (RPM) rmax: max centrifuging radius in cm by using the bottom of tubes 3.1.5 Removing the rotor Untighten the rotor fixing nut completely (screw over the stiff point) and lift the rotor vertically out of the centrifuge. (See figure 9 and 10). 3.2 Lid control 3.2.1 Lid open After the run, when the lid of the centrifuge is closed, the word "close" (M1) appears in the display "RPM | RCF" (A-1). Additionally, if there is a rotor in the centrifuge, the word "rotor"(M3) appears as well as the code number of the respective rotor, which is in the centrifuge system "71" (M4). If there is no rotor in the centrifuge, the word "rotor" (M3) flashes and additionally the word "no" (M4) appears. By pressing the key "Door Open" (7) you can release the lid of the centrifuge. As soon as the electromagnetic lid is completely released, the word "open" (M2) appears. Now you can open the lid of the centrifuge. Please refer to figure 15 below for reference. EN-8 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R M2 M4 M3 M1 A-1 1 7 Figure. 15 During the run you can call up the rotor type at any time by pressing the key "Door Open" (7). 3.2.2 Lid lock The lid should only be put down slightly. An electromagnetic lid lock closes the lid, at the same time the word "open" (M2) disappears (refer to figure 15). As a sign that the centrifuge is ready for starting, in the display "RPM | RCF" (A-1) the word "close" (M1) appears. Simultaneously the word "rotor" (M3) is displayed, as well as the code number of the rotor, which is in the centrifuge system, "no 71" (M4). With that, all rotor specific data, like e. g. max. speed, acceleration etc., are adopted. ATTENTION: Don't grip your fingers between lid and device or locking mechanism when closing the lid! 3.3 Preselection 3.3.1 Preselection of speed / RCF-value This pre-selection is activated through the key "RPM | RCF" (4) (refer to figure 16 below). By pressing the key once the word "rpm" (M5) flashes. By pressing the key twice the pre-selection of the centrifugal forces can be selected. Then the flashing word "rcf" (M6) appears. You can set the desired values with the adjusting knob (1). In the display (A-1) the regulated value is shown permanently, before, during and after the run. M5 M6 A-1 1 4 5 Figure. 16 As long as no rotor is inserted, the speed is adjustable between 200 rpm and maximum revolution of the centrifuge. If there is a rotor in the centrifuge the speed can only be pre-selected until the maximum permissible revolution of that rotor. It is the same with the pre-selection of the RCF-value. The setting range is between the centrifugal force at 200 rpm and the maximum permissible centrifugal force of the rotor. See "Table 3: max. speed and RCF-values for permissible rotor" (See APPENDIX). All important values are listed there. FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-9 ATTENTION: Please also check the maximum permissible revolutions of your test tubes with the manufacturer. 3.3.2 Preselection of running time The running time can be pre-selected in three different ranges from 10 seconds up to 99 hours 59 minutes. 1. Range from 10 seconds up to 59 minutes 50 seconds in steps of 10 seconds 2. Range from 1 hour up to 99 hours 59 minutes in steps of 1 minute. 3. The continuous run "cont", which can be interrupted by the key "Stop"(10) (refer to figure 17). The running time can be pre-selected with the lid open or closed. To activate the setting of the running time press the key "Time" (6). In the display "Time/Prog" (A-3) flashes the indication "m : s" or "h : m", depending on the previous setting. To set the desired value, use the adjusting knob (1). After exceeding 59 min 50 sec the indication changes automatically into "h : m". After exceeding 99 hours 59 min the word "cont" appears in the display "Time/Prog" (A-3). That continuous run can only be interrupted by pressing the key "Stop" (10). The time countdown starts as soon as the set speed is reached. The display always shows the remaining running time. (See figure 17) A-3 1 10 6 Figure. 17 3.3.3 Preselection of brake intensity and acceleration This function is activated through the key "Accel/Decel" (5) (refer to figure18). By pressing the key once the word "accel" (M7) flashes in the display "Acc/Dec" (A-2). The desired acceleration can be pre-selected by the adjusting knob (1). The value 0 is equivalent to the lowest and the value 9 to the highest acceleration. By pressing the key "Accel/Decel" (5) twice, the display "Acc/Dec" (A-2) indicates the word "decel"(M8). Now the desired brake intensity can be pre-selected by the adjusting knob (1). The value 9 is equivalent to the shortest and the value 0 to longest possible brake time. See "table 5: acceleration and deceleration times" (APPENDIX). There the acceleration and deceleration times for the acceleration and deceleration stages 0 to 9 for permissible rotors are shown. M7 M8 A-2 1 5 Figure. 18 EN-10 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 3.3.4 Pre-selection of temperature (Only Refrigerated Models) This function is activated by the key "Temp/Setup" (13). After pressing this key in the display "Time/Prog" the indication "°C " (A-4) flashes. By the adjusting knob (1) the desired test temperature can be pre-selected in steps of 1°C in a range from -20°C up to +40°C. The value is indicated permanently in the display (figure 19) - before, during and after the run. Please notice the respective lowest temperatures of the rotors at maximum speed! A-4 1 13 Figure. 19 3.3.5 Pre-cooling (Only Refrigerated Models) If the samples are temperature-sensitive it is useful to pre-cool the centrifuge, the rotor and eventually the buckets to the required working temperature. Therefore, insert the desired rotor and pre-set the respective temperature. By simultaneous pressing the keys "Temp/Setup" (13) (refer to figure 20) and "Time" (6) you start the run. While running, the unit chooses automatically a rotational speed that is equivalent to 20 % of the permit ted rotational speed of the respective rotor. After the pre-set temperature is reached you can leave the pre-cooling run with the "Stop" key (10). Depending on the inserted rotor the pre-cooling goes between approx. 10 and 20 min. A-4 13 10 6 Figure. 20 3.4 Radius correction If you use adapters or reducers it could change the centrifugal radius of the respective rotor. In that case you can correct the radius manually. Please proceed as follows: Close the lid, then press the key "Time" (6) (refer to figure 21) and the key "Prog/Setup" (11) at the same time and hold them. In the display "Time/Prog" (A-3) the word "radius" (M9) appears. By the adjusting knob (1) you can preselect the respective radius correction (See Table 7, APPENDIX) in steps of 0.1 cm. As soon as you have set a radius correction the word "radius" (M9) appears. This word will be visible until you put the radius correction back to 0 again. FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-11 A-3 M9 11 1 6 Figure. 21 3.5 Program 3.5.1 Storage of programs You can store up to 99 runs with all relevant parameters, including the used rotors. You can use any free program number and call it up again. Put the needed rotor into the centrifuge. By pressing the key "Prog/Setup" (11) in the display "Time/Prog" (A-3) the word "programm" appears. With the adjusting knob (1) you can chose the desired program number. If a program number is already occupied, in the display "RPM | RCF" (A-1), the words "rotor" (M3) and "xx" (M4) will appear. In case of free program numbers, 0 appears. M4 M3 A-3 A-1 M10 11 1 12 Figure. 22 Close the lid of the centrifuge. Now proceed as described previously to set all important run parameters. If the lid isn't closed when storing the program, in the display "RPM/RCF" (A-1), the words "FirSt" and "CLOSE Lid" (See figure 23) flashes alternately. If you want to start the run without storing the program, in the display "RPM/RCF" (A1) ,the words "First" and "PrESS StoreE" (See figure 24) flashes alternately. Figure. 23 Figure. 24 EN-12 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R For adaption of data press the key "Store" (12) (refer to figures 23 and 24) for approx. 1 second. If the program is stored correctly, the word "StorE" appears in the display "RPM/RCF" (A-1). As a result, the word "programm" (M10) disappears. As soon as the key "Store" (12) is released, the word "programm xx" (M10) reappears – the (xx) stands for the chosen program location. If all program numbers are occupied you can take an old number that is not necessary anymore and just put in the new parameters. 3.5.2 Recall of stored programs To recall stored programs press the key "Prog/Setup" (11) (refer to figure 25) while the lid is already closed. Inside the display "Time/Prog" (A-3), "programm –"(M10) appears. The desired program number can be pre-selected with the adjusting knob (1). In the respective displays the stored values for that program will appear. If the wrong rotor is inside the centrifuge for the pre-selected program, in the display "RPM | RCF" (A-1), the word "rotor" (M3) flashes. At the same time the word "FALSE" and the stored rotor number " xx" (M4) will flashing by turns. M3 M4 A-3 A-1 M10 11 1 12 Figure. 25 3.5.3 Leaving program mode To leave the program mode just press the key "Prog/Setup" (11) (refer to figure 25). Then inside the display "Time/Prog" the word "programm" appears. Set the display to "programm--" (M10) with the adjusting knob (1). 3.6 Starting and stopping the centrifuge 3.6.1 Starting the centrifuge You can start the centrifuge either with the "Start" key (9) (refer to figure 26) or the "Quick Spin" key (8). By the "Start" key (9) you can start stored runs or runs with manually pre-selected parameters. When the respective pre-selected running time has ended the centrifuge will stop automatically. By the "Quick Spin" key (8) you can start runs, which will last just a few seconds. By pressing the "Quick Spin" key (8) the centrifuge accelerates up to the pre-selected revolution. In the display "Time/Prog" (A-3) the passed running time is indicated from the date of pressing the "Quick Spin" key (8). By releasing the "Quick Spin" key (8) the centrifuge stops and the running time is indicated until the opening of the lid. FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-13 A-3 9 1 8 Figure. 26 3.6.2 The "STOP" key By the "Stop" key (10) (See figure 27) you can interrupt the run at any time. After pressing the key the centrifuge decelerates with the respective pre-selected intensity down to stand still. 10 Figure. 27 3.7 Imbalance detection In case of the rotor not being equally loaded, the drive will turn off during acceleration. The rotor decelerates to stand still. When in the display "Time/Prog" (A-3) the word "error” (M11) together with the number "01” appear, the weight difference of the samples is too large. Distribute the weight evenly. Load the rotor as described in chapter 3.1.2 and 3.1.3. When inside the display "Time/Prog" (A-3) the word "error” together with the number "02" (See figure 28) appear, it could be due to the following reason: The imbalance switch is defective. A-3 M11 Figure. 28 EN-14 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4. SETTING 4.1 Basic adjustments 4.1.1 Access to mode "Operating Data" When using the centrifuge, the following parameters can be set: - Temperature indication °C or °F - Acoustic signal turn on/off - Keyboard sound turn on/off - Volume pre-selection of sound signal - Song selection of sound signal "end of run" While the centrifuge is turned off, press simultaneously the keys "Time"(6) and "Door Open" (7) and turn on the main switch of the centrifuge. Now release both keys and as a result a display test is executed for approx. 5 seconds. All indicators will appear at the same time (See figure 29). 7 6 Figure. 29 ATTENTION: Please notice that you must enter the program as described under point 4.1.1 to change the adjustments of the points 4.1.2 - 4.1.7. A-1After you have stored the settings you can change to normal program mode again by switching off the centrifuge for a short while. All changed settings must be confirmed by the key "Start"(9). The word "Store"(12) appears in the display "RPM | RCF"(A-1) - Only then the pre-selections are valid! 12 9 Figure. 30 4.1.2 Temperature indication Proceed as described under point 4.1.1 to enter this program mode and then press the key "Accel/Decel" (5). In the display "Acc/Dec" (A-2) the word "Service" appears. Now select the letter "C" with the adjusting knob (1). As a result, in the display "RPM | RCF" (A-1), the words "CELSI/temp" appear. If you press the key "RPM | RCF" (4), FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-15 the word "CELSI" flashes and you can change the display into Fahrenheit "FAREN", with the adjusting knob (1) (See figure 31). After you have stored the settings (See 4.1.1) you change back to the normal program mode again by switching off the centrifuge for a short while. 1 4 5 Figure. 31 4.1.3 Signal turn on / off Proceed as described under point 4.1.1 to enter this program mode and then press the key "Accel/Decel" (5). In the display "Acc/Dec" (A-2) the word "Service" flashes. Now select the letter "L" with the adjusting knob (1). As a result, the words "On Sound" appears in the display "RPM | RCF" (4). If you press the key "RPM | RCF" (4) now, the word "On" flashes and you can switch off the sound with the adjusting knob (1) (See figure 32). After you have stored the settings (See 4.1.1) you change back to the normal program mode again by switching off the centrifuge for a short while. A-1 A-2 1 4 5 Figure. 32 4.1.4 Volume pre-selection of sound signal Proceed as described under point 4.1.1to enter this program mode and then press the key "Accel/Decel" (5). In the display "Acc/Dec" (A-2) the word "Service" flashes. Now select the letter "U" with the adjusting knob (1). As a result, in the display "RPM | RCF" (A-1) the words "Vol=0- 9/Sound" appear. After pressing the key "RPM | RCF" (4), you can adjust the desired volume between 0 (low) and 9 (loud) with the adjusting knob (1) (See figure 33). After you have stored the settings (see 4.1.1) you can change back to the normal program mode again by switching off the centrifuge for a short period. A-1 A-2 11 4 5 Figure. 33 EN-16 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4.1.5 Song selection for sound signal - end of run Proceed as described under point 4.1.1to enter this program mode and then press the key "Accel/Decel" (5). In the display "Acc/Dec" (A-2) the word "Service" flashes. Now select the letter "G". with the adjusting knob (1). As a result, in the display "RPM | RCF" (A-1), the word "SonGo/Sound" appears. After pressing the key "RPM | RCF" (4), you can select a song with the adjusting knob (1). (See figure 34). After you have stored the settings (see 4.1.1) you can change back to the normal program mode again by switch off the centrifuge for a short while. A-1 A-2 1 4 5 Figure. 34 4.1.6 Keyboard sound turn on / off Proceed as described under point 4.1.1to enter this program mode and then press the key "Accel/Decel" (5). In the display "Acc/Dec" (A-2) the word "Service" flashes. Now select the letter "B". with the adjusting knob (1). As a result, in the display "RPM | RCF" (A-1), the word "ON/BEEP " appears. After pressing the key "RPM | RCF" (4), you can turn the keyboard sound (On) or (Off) with the adjusting knob (1). (See figure 39). After you have stored the settings (see 4.1.1) you can change back to the normal program mode again by switch off the centrifuge for a short while. A-1 A-2 1 4 5 4.1.7 Call up operating data Figure. 35 ATTENTION: This should only be performed by advanced user or service engineer. In the mode "Basic Adjustments" you can call up the operating data of the centrifuge. Please proceed as described under point 4.1.2 to enter this program mode. Press the key "Accel/Decel" (5). In the display "Acc/Dec" (A-2) the word "Service" flashes. With the adjusting knob (1) the different information can be accessed: A= previous starts of the centrifuge FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-17 H= previous operating hours S= software version r= converter software E= list of previous error messages h= running time of the motor The list of the last 99 error messages can be looked over by pressing the key "RPM | RCF" (4) and scroll through it by the adjusting knob (1). The respective error codes appear in the display "RPM | RCF" (A-1). Please refer to "Table 6: error messages" (see APPENDIX). To change back to normal program mode again, switch off the centrifuge for a short period. A-1 A-2 1 4 5 Figure. 36 5. MAINTENANCE 5.1 Maintenance and cleaning 5.1.1 General Care: Maintenance of the centrifuge is confined to keeping the rotor, the rotor chamber and the rotor accessories clean as well as to regularly lubricating the rotor insert bolts of a swing out rotor (if available). The most suitable lubricant is the High TEF oil. Lubricants containing molycote and graphite are not allowed. Please pay special attention to anodized aluminum parts. Breakage of rotors can be caused even by slight damage. In case of rotors, buckets or tube racks getting in touch with corrosive substances the concerned spots have to be cleaned carefully. Corrosive substances are for instance: alkalis, alkaline soap solutions, alkaline amines, concentrated acids, solutions containing heavy metals, water-free chlorinated solvents and saline solutions. e.g. salt water, phenol, halogenated hydrocarbons. Cleaning – units, rotors, accessories: − − − − Turn the device off and disconnect it from the power supply before you begin any cleaning or disinfecting. Do not pour liquids into the housing interior. Do not spray disinfectant on the device. Thorough cleaning not only has its purpose in hygiene but also in avoiding corrosion due to pollution. In order to avoid damaging anodized parts such as rotors, reduction plates etc., only pH-neutral Detergents with a pH-value of 6-8 may be used for cleaning. Alkaline cleaning agents (pH-value > 8) must not be used. EN-18 − − − FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R After cleaning, please ensure all parts are dried thoroughly, either by hand or in a hot-air cabinet (max.Temperature + 50°C). It is necessary to coat anodized aluminum parts with anti-corrosion oil regularly in order to increase their lifespans and reduce corrosion predisposition. Due to humidity or not hermetically sealed samples, condensate may be formed. The condensate has to be removed from the rotor chamber with a soft cloth regularly. The maintenance procedure has to be repeated every 10 to 15 runs, or at least once a week. - Connect the unit to the power supply, after the equipment is completely dry. Do not carry out disinfection with UV-, beta- and gamma-rays or other high energy radiation. Metal rotors can be autoclaved. Rotor lid and adapters can also be autoclaved (max. 121°C, 20 min). The tube racks are made of PP and cannot be autoclaved at 134°C. 5.1.2 Cleaning and disinfection of the unit 1. Open the lid before you turn off the unit. Disconnect it from the power supply. 2. Open the rotor nut by turning the rotor key counter clockwise. 3. Remove the rotor. 4. For cleaning and disinfection of the unit and the rotor chamber us e the above mentioned cleaner. 5. Clean all accessible areas of the device and its accessories, including the power cord with a damp cloth. 6. Wash the rubber seals and rotor chamber thoroughly with water. 7. Rub the dry rubber seals with glycerol or talc to prevent these to becoming brittle. Other components of the unit, e.g. the lid lock, motor shaft and rotor must not be greased. 8. Dry the motor shaft with a soft, dry and lint-free cloth. 9. Control the unit and accessories for damage. Make sure that the centrifuge is turned off the unit and disconnect the unit from the power supply. Then remove adherent dust from the ventilation slots in the centrifuge by using a soft brush. Do this at least every six months. 5.1.3 Cleaning and disinfection of the rotor 1. Clean and disinfect the rotors, rotor lids and adapters with the above mentioned cleaner. 2. Use a bottle brush to clean and disinfect the rotor bores. 3. Rinse the rotors, rotor lid and adapter with clear water. Particularly the drillings of angle rotors. 4. For drying of the rotors and accessories set them on a towel. Place the angle rotors with bores down. 5. Dry the rotor cone with a soft, dry and lint-free cloth and look for damage. Do not grease the rotor cone. 5.1.4 Disinfection of aluminum rotors In case of infectious material spilling into the centrifuge, the rotor and rotor chamber have to be disinfected directly after the run. Rotors may be autoclaved at a maximum temperature of 121°C. 5.1.5 Disinfection of PP-rotors Autoclaving The recommended time for autoclaving: 15 – 20 min at 121°C (1 bar) ATTENTION: The sterilization time of 20 min. must not be exceeded. Repeated sterilization will cause reduction of the mechanical resistance of the plastic material Before autoclaving the PP-rotor and adapter must be thoroughly cleaned to avoid the burning in of dirty residues. You can disregard the consequences of some chemical residues to plastic materials at ambient temperatures. But at the high temperatures during autoclaving those residues may corrode and destroy the plastic. The objects must be thoroughly rinsed with distilled water after the cleaning but before the autoclaving. Residues of any cleaning liquids may cause fissures, whitening and stains. Gas sterilization Adapters, bottles and rotors may be gas sterilized with Ethylenoxyd. Make sure to air out the items after the sterilization and before using them again. ATTENTION: Because the temperature may rise during the sterilization, rotors, adapters and bottles must not be closed and must be totally unscrewed FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-19 Chemical sterilization Bottles, adapters and rotors may be treated with the usual liquid disinfectants. ATTENTION: Before applying any other cleaning or decontamination method than recommended by the manufacturer, contact the manufacturer to ensure that it will not damage the unit or the rotor. 5.1.6 Glass breakage With high g-values, the rate of glass tube breakage increases. Glass splinters have to be removed immediately from rotor, buckets, adapters and the rotor chamber itself. Fine glass splinters will scratch and therefore damage the protective surface coating of a rotor. If glass splinters remain in the rotor chamber, fine metal dust will build up due to air circulation. This very fine, black metal dust will significantly pollute the rotor chamber, the rotor, the buckets and the samples. If necessary, replace the adapters, tubes and accessories to avoid further damage. Check the rotor bores regularly for residues and damage. ATTENTION: Please check the relevant specifications of the tubes ce ntrifuges with the manufacturer. 5.2 Life time of rotors, buckets, accessories Rotors and rotor lid made of aluminum or stainless steel, have an operating time of max. 7 years from first use. Transparent rotor lids and caps made of PC or PP as well as rotors, tube racks and adapters of PP have a maximum operating time up to 3 years from first use. Condition for the operating time: Proper use damage-free condition, recommended care. 6. TROUBLESHOOTING 6.1 Error message: Cause / Solution The error messages are listed to help localize possible errors faster. The diagnosing referred to in this chapter may not always be the case, as they are only theoretically occurring errors and solutions. Please keep us informed about any kind of error occurring, which is not listed in this chapter. Only through your information are we able to improve this operation manual. Many thanks in advance for your support. 6.2 Survey of possible error messages and their solutions 6.2.1 Lid release during power failure (Emergency Lid Release) In case of power failure or malfunction, the lid of the centrifuge can be opened manually in order to retrieve your samples. For FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (motor driven lock) Please proceed as follows: 1. Switch off the centrifuge and unplug the power cord, wait until the rotor has come to a standstill (this may take several minutes) 2. On the left side of the centrifuge housing there is a plastic stopper. Remove this stopper and behind it there is a hexagon nut. 3. Take the included box spanner, put it in the hole and lock the box spanner with the hexagon nut (See figure 37). 4. Now turn the box spanner to the right side (clockwise) up to the limit. ATTENTION: 3 a) Just turn to the limit, don´t tighten the nut. b) Now open the lid of the centrifuge. c) Switch the centrifuge on again, to resume work. EN-20 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figure. 37 For 5714 Please proceed as follows (see Figure 38): ATTENTION: • Switch the centrifuge off and unplug the power cord, wait until the rotor stands still (this may take several minutes). At the right side of the centrifuge there is a plastic stopper (Figure 38). Remove this stopper, which is connected to the lid lock, horizontally from the housing until the centrifuge lid opens. • Now open the lid of the centrifuge Figure. 38 6.2.2 Description of the error message system The error message "error" (M11) is shown in the "Time/Prog" (A-3) display (See figure 39). Detailed information about possible error messages are in "table 6: error messages" (See Appendix). A-3 M11 Figure. 39 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 7. RECEIPT OF CENTRIFUGES TO REPAIR Health risk from contaminated equipment, rotors and accessories. In case of returning the centrifuge for repairing to the manufacturer, please notice the following: The centrifuge must be decontaminated and cleaned before the shipment for the protection of persons, environment and material. Decontamination certificate at goods return delivery (See APPENDIX) We reserve the right to not accept contaminated centrifuges. Further on all costs occurred for the cleaning and disinfection of the units will go to the debit of the customer’s account. 8. TRANSPORT, STORAGE AND DISPOSAL 8.1 Transport Before transporting, take out the rotor. Only transport the unit in the original packaging. Install the transport protection material to secure the motor shaft, when transporting over longer distances. Air temperature rel. humidity General transportation -25 to 60 °C 10 to 75 % Air pressure 30 to 106 kPa 8.2 Storage During storage of the centrifuge the following environmental conditions must be observed: Air temperature rel. Humidity Air pressure -25 to 55 °C 10 to 75 % 70 to 106 kPa in transport packaging 8.3 Transporting, Installing, Transferring and Disposing of the Centrifuge FC5714/FC5718/ FC5718R/FC5816/FC5816R These instructions complement the previous general instructions in chapter 8 and do not replace them. 8.3.1 Transport • Please transport the device in the original packaging. • The centrifuge should always be transported using a mechanical transport device. 8.3.2 Installation  Opening the carton and lifting out the device. 1. Cut the adhesive tape. 2. Open all 4 flaps of the carton. 3. Remove the accessories. 4. Carefully lift the centrifuge from the carton. WARNING: Lifting Hazard. Single person lift could cause injury. Use a mechanical lifting device or team lifting procedures when lifting or moving the equipment. EN-21 EN-22 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R  Place the device on a stable, horizontal and non-resonant lab bench 1. Remove the front and back transport protection material. 2. Remove the plastic sleeve. 3. Observe a minimum distance of 30 cm to adjoining devices at the sides and from the rear side to the wall. 4. Install the device in a well-ventilated location which is protected from direct sunlight to prevent it from overheating.  Connect the device 1. After installation, wait for four hours before switching the centrifuge on in order to avoid damage to the compressor. 2. Check that the mains voltage and frequency match the requirements on the device name plate(see rear side of the device) and then connect the device to the power supply.  Remove the transport protection material from the rotor chamber 1. Switch on the device at the mains power switch. 2. Open the centrifuge lid using the open button. 3. Remove the transport protection material. 4. Place the rotor vertically onto the motor shaft. 5. Turn the rotor nut using the rotor key clockwise until the rotor nut is tightened.  The device is now ready to use Retain the packaging and all transport protection material for shipping the device at a later date. 8.3.3 Packing Pack the centrifuge in reverse order. 8.3.4 Passing on the Device When passing the equipment on to third parties, please make sure to also include this instruction manual. FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-23 9. TECHNICAL DATA 9.1 Specifications 9.1.1 Centrifuge FC5714 Model Speed Range FC5714 200 rpm -14000 rpm;10 rpm/set Maximum RCF Maximum Capacity(Rotor) 18624 x g;10 x g/set 4x100ml Temperature range(N/A) Running Time Noise level (depending on the rotor) Air cool 10 sec to 99 hr 59 min 59 sec or continuous ≤ 63 ± 2 dB(A) Allowable density at maximum speed Allowable kinetic energy Mains power connection AC 1.2 g/ml 5595 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz Voltage fluctuation Current consumption ± 10 % 1.3 A ± 10 % 2.4 A Power consumption 240 W 300 W 362 x 493 x 330 mm 14.3 x 19.4 x 13.0 in Dimensions (W × D × H) 30 kg 66 lb 580 x 490 x 460 mm Net Weight (without rotor) Shipping Dimensions (W × D × H) 22.8 x 19.3 x 18.1 in 32.5 kg Shipping Weight (without rotor) 72 lb Ambient conditions (EN/IEC 61010-1) Environment For indoor use only Altitude Ambient temperature Use up to an altitude of 2000 m 2°C up to 35 °C Max. relative humidity Max. relative humidity 80 % for temperatures up to 31°C, decreasing linearly to 50 % relative humidity up to 35°C. Overvoltage category (IEC 60364-4-443) Degree of contamination II 2 Class of protection Not suitable for use in hazardous environments. I EN-24 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.2 Centrifuge FC5718 Model Speed Range FC5718 200 rpm -18000 rpm;10 rpm/set Maximum RCF Maximum Capacity (Rotor) Temperature range (N/A) 23542 x g;10 x g/set 4x100ml Air cool Running Time Noise level Noise level (depending on the rotor) Allowable density at maximum speed Allowable kinetic energy Mains power connection AC 10 sec to 99 hr 59 min 59 sec or continuous ≤ 60 ± 2 dB(A) 1.2 g/ml 230 V ~ 50/60 Hz Voltage fluctuation Current consumption Power consumption ± 10 % 2.0 A 455 W Dimensions (W × D × H) Net Weight (without rotor) Shipping Weight (without rotor) 117 lb Ambient conditions (EN/IEC 61010-1) Environment Max. relative humidity Overvoltage category (IEC 60364-4-443) Degree of contamination ± 10 % 4.0 A 475 W 408 x 499 x 351 mm 16.1 x 19.7 x 13.8 in 43 kg 95 lb 650 x 520 x 490 mm 25.6 x 20.5 x 19.3 in 53 kg Shipping Dimensions (W × D × H) Altitude Ambient temperature 16672 Nm 120 V ~ 50/60 Hz for indoor use only Use up to an altitude of 2000 m 2°C up to 35 °C Max. relative humidity 80 % for temperatures up to 31°C, decreasing linearly to 50 % relative humidity up to 35°C. II 2 Class of protection Not suitable for use in hazardous environments. I FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-25 9.1.3 Centrifuge FC5718R Model Speed Range Maximum RCF FC5718R 200 rpm -18000 rpm;10 rpm/set 23542 x g;10 x g/set Maximum Capacity(Rotor) Temperature range(Digital) Running Time 4x100ml -20~40℃ 10 sec to 99 hr 59 min 59 sec or continuous Noise level (depending on the rotor) Allowable density at maximum speed Allowable kinetic energy Mains power connection AC Voltage fluctuation ≤ 60 ± 2 dB(A) 1.2 g/ml 25111 Nm 230 V ~ 50/60 Hz ± 10 % 120 V ~ 50/60 Hz ± 10 % 3.0 A 660 W 6.0 A 660 W Current consumption Power consumption 407 x 731 x 359 mm Dimensions (W × D × H) 16.0 x 28.8 x 14.1 in 60 kg Net Weight (without rotor) 132 lb 840 x 640 x 590 mm 33.1 x 25.2 x 23.2 in Shipping Dimensions (W × D × H) 77 kg 170 lb Shipping Weight (without rotor) Ambient conditions (EN/IEC 61010-1) Environment Altitude Ambient temperature Max. relative humidity Overvoltage category (IEC 60364-4-443) for indoor use only Use up to an altitude of 2000 m 2°C up to 35 °C Max. relative humidity 80 % for temperatures up to 31°C, decreasing linearly to 50 % relative humidity up to 35°C. II Degree of contamination Class of protection Not suitable for use in hazardous environments. 2 I EN-26 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.4 Centrifuge FC5816 Model Speed Range Maximum RCF FC5816 200 rpm -15000 rpm;10 rpm/set 21379 x g;10 x g/set Maximum Capacity(Rotor) Temperature range(N/A) Running Time 6 x 250ml Air cool 10 sec to 99 hr 59 min 59 sec or continuous Noise level (depending on the rotor) Allowable density at maximum speed ≤ 61 ± 2 dB(A) 1.2 g/ml Allowable kinetic energy Mains power connection AC Voltage fluctuation 34363 Nm Current consumption Power consumption 230 V ~ 50/60 Hz ± 10 % 120 V ~ 50/60 Hz ± 10 % 2.4 A 530 W 4.2 A 520 W 446 x 538 x 354 mm Dimensions (W × D × H) 17.6 x 21.2 x 13.9 in 52 kg Net Weight (without rotor) 115 lb 840 x 640 x 590 mm 33.1 x 25.2 x 23.2 in Shipping Dimensions (W × D × H) 77 kg 170 lb Shipping Weight (without rotor) Ambient conditions (EN/IEC 61010-1) Environment Altitude for indoor use only Use up to an altitude of 2000 m Ambient temperature 2°C up to 35 °C Max. relative humidity Max. relative humidity 80 % for temperatures up to 31°C, decreasing linearly to 50 % relative humidity up to 35°C. Overvoltage category (IEC 60364-4-443) Degree of contamination II 2 Class of protection Not suitable for use in hazardous environments. I FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-27 9.1.5 Centrifuge FC5816R Model Speed Range Maximum RCF FC5816R 200 rpm -16000 rpm;10 rpm/set 24325 x g;10 x g/set Maximum Capacity(Rotor) Temperature range(Digital) Running Time 6 x 250ml -20~40℃ 10 sec to 99 hr 59 min 59 sec or continuous Noise level (depending on the rotor) Allowable density at maximum speed ≤ 63 ± 2 dB(A) 1.2 g/ml Allowable kinetic energy Mains power connection AC Voltage fluctuation 34363 Nm Current consumption Power consumption 230 V ~ 50/60 Hz ± 10 % 120 V ~ 50/60 Hz ± 10 % 3.7 A 785 W 7.8 A 850 W 723 x 538 x 354 mm Dimensions (W × D × H) 28.5 x 21.2 x 13.9 in 77 kg Net Weight (without rotor) 170 lb 840 x 640 x 590 mm 33.1 x 25.2 x 23.2 in Shipping Dimensions (W × D × H) 87 kg 192 lb Shipping Weight (without rotor) Ambient conditions (EN/IEC 61010-1) Environment Altitude for indoor use only Use up to an altitude of 2000 m Ambient temperature 2°C up to 35 °C Max. relative humidity Max. relative humidity 80 % for temperatures up to 31°C, decreasing linearly to 50 % relative humidity up to 35°C. Overvoltage category (IEC 60364-4-443) Degree of contamination II 2 Class of protection Not suitable for use in hazardous environments. I EN-28 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.6 Centrifuge FC5916 Model Speed Range Maximum RCF Maximum Capacity(Rotor) Temperature range(N/A) FC5916 200 rpm -16000 rpm;10 rpm/set 24325 x g;10 x g/set 4x750ml Air cool Running Time Noise level (depending on the rotor) Allowable density at maximum speed Allowable kinetic energy Mains power connection AC Voltage fluctuation Current consumption Power consumption 10 sec to 99 hr 59 min 59 sec or continuous 230 V ~ 50/60 Hz ± 10 % 2.8 A 640 W Dimensions (W × D × H) Net Weight (without rotor) Shipping Dimensions (W × D × H) Shipping Weight (without rotor) Ambient conditions (EN/IEC 61010-1) Environment Altitude Ambient temperature ≤ 63 ± 2 dB(A) 1.2 g/ml 60629 Nm 120 V ~ 50/60 Hz ± 10 % 5.6 A 680 W 540 x 670 x 390 mm 21.3 x 26.4 x 15.4 in 85 kg 187 lb 780 x 670 x 590 mm 30.7 x 26.4 x 23.3 in 98 kg 216 lb For indoor use only Use up to an altitude of 2000 m 2°C up to 35 °C Max. relative humidity 80 % for temperatures up to 31°C, Max. relative humidity decreasing linearly to 50 % relative humidity up to 35°C. Overvoltage category (IEC 60364-4-443) Degree of contamination Class of protection Not suitable for use in hazardous environments. II 2 I FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-29 9.1.7 Centrifuge FC5916R Model Speed Range Maximum RCF Maximum Capacity(Rotor) FC5916R 200 rpm -16000 rpm;10 rpm/set 26331 x g;10 x g/set 4x750ml Temperature range(Digital) -20~40℃ Running Time Noise level (depending on the rotor) Allowable density at maximum speed Allowable kinetic energy Mains power connection AC Voltage fluctuation Current consumption Power consumption 10 sec to 99 hr 59 min 59 sec or continuous 230 V ~ 50/60 Hz ± 10 % 7.2 A 1630 W Dimensions (W × D × H) Net Weight (without rotor) Shipping Dimensions (W × D × H) Shipping Weight (without rotor) ≤ 63 ± 2 dB(A) 1.2 g/ml 54458 Nm 120 V ~ 50/60 Hz ± 10 % 20 A 1750 W 730 x 670 x 390 mm 28.7 x 26.4 x 15.4 in 118 kg 260 lb 900 x 750 x 560 mm 40.0 x 29.5 x 22.0 in 137 kg 137 kg Ambient conditions (EN/IEC 61010-1) Environment Altitude Ambient temperature for indoor use only Use up to an altitude of 2000 m 2°C up to 35 °C Max. relative humidity 80 % for temperatures up to 31°C, Max. relative humidity decreasing linearly to 50 % relative humidity up to 35°C. Overvoltage category (IEC 60364-4-443) Degree of contamination II 2 Class of protection I Not suitable for use in hazardous environments. EN-30 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.2 Drawings and dimensions D H W D H W FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Model FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R 10. W (mm / in.) 362 / 14.3 408 / 16.1 408 / 16.1 446 / 17.6 723 / 28.5 540 / 21.3 730 / 28.7 D (mm / in.) 493 / 19.4 499 / 19.7 731 / 28.8 538 / 21.2 538 / 21.2 670 / 26.4 670 / 26.4 H (mm / in.) 330 / 13.0 351 / 13.8 359 / 14.1 354 / 13.9 354 / 13.9 390 / 15.4 390 / 15.4 COMPLIANCE Compliance to the following standards is indicated by the corresponding mark on the product. Marking Standard This product complies with the applicable harmonized standards of EU Directives 2011/65/EU (RoHS), 2014/30/EU (EMC) and 2014/35/EU (LVD) and 2014/31/EU (NAWI). The EU Declaration of Conformity is available online at www.ohaus.com/ce. Disposal In conformance with the European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. For disposal instructions in Europe, refer to www.ohaus.com/weee. Thank you for your contribution to environmental protection. EN-31 EN-32 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FCC Supplier Declaration of Conformity Unintentional Radiator per 47CFR Part B Trade Name: OHAUS CORPORATION Model: FC... Party issuing Supplier’s Declaration of Conformity: Ohaus Instruments (Changzhou) Co., Ltd. 2F, 22 Block, 538 West Hehai Road, Xinbei District, Changzhou Jiangsu 213022 China Phone: +86 519 85287270 Responsible Party – U.S. Contact Information: Ohaus Corporation 7 Campus Drive, Suite 310 Parsippany, NJ 07054 United States Phone: +1 973 377 9000 Web: www.ohaus.com FCC Compliance Statement: Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: —Reorient or relocate the receiving antenna. —Increase the separation between the equipment and receiver. —Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. —Consult the dealer or an experienced radio/TV technician for help. Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11. APPENDIX TABLE 1: PERMISSIBLE NET WEIGHT TABLE 2: LOWEST TEMPERATURES AT MAX. SPEED TABLE 3: MAX. SPEED AND RCF-VALUES FOR PERMISSIBLE ROTORS TABLE 4: ACCELERATION AND DECELERATION TIMES TABLE 5: ERROR MESSAGES TABLE 6 (PART 1): RADIUS CORRECTION TABLE 7: REDEMPTION FORM / DECONTAMINATION CERTIFICATE EN-33 EN-34 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.1 Table 1: Permissible net weight FC5916R FC5916 FC5816R FC5816 Permissible net weight FC5718R Rotor Description FC5718 Rotor number FC5714 Max Speed ( x1000 rpm) 14.0 14.0 15.0 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 24x 3.4 g 15.0 16.0 16.0 16.0 30314820 Rotor, Sw ing out, 4x 250ml, ID 4.5 4x 536 g 4.5 8.0 30314821 Rotor, Angle, 6x 250ml FB, ID 8.0 6x 355 g 8.0 8.0 4.5 30314822 Rotor, Sw ing out, 4x 100ml, ID 4x 340 g 4.5 4.5 4.0 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 4x 465 g 5.0 5.0 4.5 4.5 4.5 30314824 Rotor, Sw ing out, 2x 3MTP, ID 2x 310 g 4.5 4.5 4.5 4.5 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 6x 140 g 11.0 13.5 9.0 9.0 30314826 Rotor, Angle, 6x 85ml RB, ID 11.0 6x 140 g 13.0 11.0 11.0 12.0 12.0 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 4x 140 g 12.0 12.0 15.0 15.0 30314828 Rotor, Sw ing out, 16x 50ml, ID 4x 557 g 4.5 4.5 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-35 7.5 7.5 30314829 Rotor, Angle, 10x 50ml FA, ID 9.0 10x 74 g 10.5 10.0 10.0 6.0 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 6x 72 g 6.0 6.0 12.0 12.0 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 13.0 6x 94 g 13.0 13.0 13.0 4.5 4.5 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30x 32 g 4.5 4.5 4.5 12.0 12.0 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 12.0 20x 18 g 12.0 12.0 12.0 6.0 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 12x 25 g 6.0 6.0 13.5 13.5 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 15.0 44x 3.8 g 16.0 16.0 16.0 12.0 13.0 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30x 3.4 g 14.0 15.0 15.0 14.0 14.0 15.0 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 24x 3.4 g 15.0 16.0 15.0 15.0 30314839 30314840 Rotor, Angle, 12x 1.5/2.0ml, ID Rotor, Angle, 64x 0.5ml, ID 12x 3.4 g 64x 2.3 g 18.0 18.0 13.5 13.5 15.0 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 4x 3.5 g 15.0 15.0 15.0 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID 4x 995 g 4x 708 g 4.0 4.0 8.0 8.0 EN-36 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.2 Table 2: Lowest temperatures at max. speed (Only Refrigerated Models) Rotor number Rotor Description 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 30314840 Rotor, Angle, 64x 0.5ml, ID 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID Model FC5718R FC5816R FC5916R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5916R FC5916R Max Speed 15000 rpm 15000 rpm 16000 rpm 8000 rpm 4500 rpm 4500 rpm 5000 rpm 5000 rpm 4500 rpm 4500 rpm 4500 rpm 13500 rpm 13500 rpm 9000 rpm 9000 rpm 13000 rpm 12000 rpm 12000 rpm 16000 rpm 7500 rpm 7500 rpm 10500 rpm 6000 rpm 6000 rpm 12000 rpm 12000 rpm 13000 rpm 4500 rpm 4500 rpm 12000 rpm 12000 rpm 12000 rpm 6000 rpm 6000 rpm 13500 rpm 13500 rpm 16000 rpm 14000 rpm 14000 rpm 15000 rpm 15000 rpm 15000 rpm 16000 rpm 18000 rpm 18000 rpm 13500 rpm 13500 rpm 15000 rpm 15000 rpm 15000 rpm 4500 rpm 8000 rpm All temperature indications refer to a room temperature of 23°C. By exceeding this value or direct solar radiation to the centrifuge, these values can´t be kept up. N-max 6 °C 6 °C 1 °C -1 ° C -7 ° C -7 ° C -3 ° C -3 ° C -6 ° C -6 ° C -8 ° C 17 ° C 17 ° C -5 ° C -5 ° C 8 °C 5 °C 5 °C 7 °C 9 °C 9 °C 2 °C -8 ° C -8 ° C -5 ° C -5 ° C -4 ° C -9 ° C -9 ° C 2 °C 2 °C -2 ° C -11.5 ° C -11.5 ° C 1 °C 1 °C 3 °C 9 °C 9 °C 3 °C 6 °C 6 °C 1 °C -2 ° C -2 ° C 2 °C 2 °C -1 ° C -1 ° C -2 ° C 4 °C 0 °C FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-37 11.3 Table 3: Max. speed and RCF-values for permissible rotors Rotor number 30304361 Rotor Description Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314820 Rotor, Sw ing out, 4x 250ml, ID 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 30314823 30314824 30314825 30314826 30314827 30314828 Rotor, Sw ing out, 4x 100ml, ID Rotor, Sw ing out, 4x 100ml, ID, Sealable Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID Rotor, Angle, 4x 85ml RB, ID, Hi Rotor, Sw ing out, 16x 50ml, ID Model Max Speed RPM RCF FC5714 14000 18624 x g FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5816 4500 3780 x g FC5816R 4500 3780 x g FC5816 8000 10016 x g FC5816R 8000 10016 x g FC5916 8000 10016 x g FC5916R 8000 10016 x g FC5714 4500 3350 x g FC5718 4500 3350 x g FC5718R 4500 3350 x g FC5714 4000 2486 x g FC5718 5000 3885 x g FC5718R 5000 3885 x g FC5714 4500 2716 x g FC5718 4500 2716 x g FC5718R 4500 2716 x g FC5816 4500 2716 x g FC5816R 4500 2716 x g FC5916 4500 2716 x g FC5916R 4500 2716 x g FC5718 11000 13932 x g FC5718R 13500 20984 x g FC5718 9000 10413 x g FC5718R 9000 10413 x g FC5816 11000 15555 x g FC5816R 13000 21726 x g FC5916 11000 15555 x g FC5916R 13000 21726 x g FC5718 12000 14809 x g FC5718R 12000 14809 x g FC5816 12000 14809 x g FC5816R 12000 14809 x g FC5916 15000 23140 x g FC5916R 16000 26328 x g FC5816 4500 3735 x g FC5816R 4500 3735 x g EN-38 30314829 30314830 30314831 30314832 30314833 30314834 30314835 30314836 30314838 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Rotor, Angle, 20x 10ml RB, ID, Hi Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable Rotor, Angle, 24x 1.5/2.0ml, ID FC5718 7500 8174 x g FC5718R 7500 8174 x g FC5816 9000 11771 x g FC5816R 10500 16022 x g FC5916 10000 14532 x g FC5916R 10500 16022 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 12000 13522 x g FC5718R 12000 13522 x g FC5816 13000 15869 x g FC5816R 13000 15869 x g FC5916 13000 15869 x g FC5916R 13000 15869 x g FC5714 4500 2830 x g FC5718 4500 2830 x g FC5718R 4500 2830 x g FC5816 4500 2830 x g FC5816R 4500 2830 x g FC5718 12000 15775 x g FC5718R 12000 15775 x g FC5816 12000 15775 x g FC5816R 12000 15775 x g FC5916 12000 15775 x g FC5916R 12000 15775 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 13500 17113 x g FC5718R 13500 17113 x g FC5816 15000 21128 x g FC5816R 16000 24039 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5714 12000 15131 x g FC5718 13000 17758 x g FC5718R 14000 20595 x g FC5916 15000 23643 x g FC5916R 15000 23643 x g FC5714 14000 18624 x g FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314839 30314840 30314841 30553085 30553086 Rotor, Angle, 12x 1.5/2.0ml, ID Rotor, Angle, 64x 0.5ml, ID Rotor, Angle, 4x 8-w PCR Strip, ID Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID EN-39 FC5718 18000 23542 x g FC5718R 18000 23542 x g FC5718 13500 16298 x g FC5718R 13500 16298 x g FC5718 15000 15343 x g FC5718R 15000 15343 x g FC5916 15000 15343 x g FC5916R 15000 15343 x g FC5916 4000 3434 x g FC5916R 4500 4346 x g FC5916 8000 10367 x g FC5916R 8000 10367 x g EN-40 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.4 Table 4: Acceleration and deceleration times Item No Rotor Description 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314820 Rotor, Sw ing out, 4x 250ml, ID 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314828 Rotor, Sw ing out, 16x 50ml, ID 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID Model FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R Acceleration* Time in sec Deceleration* Time in sec level 0 level 0 200 240 230 251 249 204 204 309 309 664 664 573 573 110 100 100 110 100 100 230 160 160 452 452 249 249 400 500 360 360 697 2313 463 549 300 300 506 506 448 448 311 311 380 380 753 753 480 480 101 89 90 level 9 25 23 25 25 20 21 21 34 34 130 130 66 66 14 15 15 14 15 15 25 20 20 43 43 27 27 40 60 40 40 85 70 48 69 60 60 60 60 50 50 34 34 63 62 115 115 60 60 11 13 12 230 240 420 610 635 421 421 458 458 2906 2906 1903 1903 170 150 150 170 150 150 340 360 360 616 616 488 488 960 1260 1050 1050 2313 1630 1654 1307 820 820 1745 1745 1251 1251 387 387 796 1069 2395 2395 1747 1747 206 239 463 level 9 35 16 17 26 27 30 30 36 36 92 92 84 84 19 15 15 17 15 15 26 15 15 38 38 23 23 40 50 40 40 70 76 46 67 40 40 49 49 45 45 36 36 37 36 72 72 68 68 14 11 11 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 30314840 Rotor, Angle, 64x 0.5ml, ID 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 30553085 Rotor Sw ingOut 4x 750ml D99mm ID SL 30553086 Rotor Angle 4x 500ml D69mm ID FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5916 FC5916R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5718 FC5718R FC5916 FC5916R FC5916 FC5916R FC5916 FC5916R *Note: accelerates from 0 to Vmax; decelerates from Vmax to 0. 360 360 446 446 264 264 157 160 160 149 149 360 360 512 512 305 305 102 88 88 196 196 257 274 233 233 250 200 210 221 221 200 210 230 251 249 204 204 240 240 150 150 130 130 100 100 483 551 575 575 EN-41 40 40 48 48 28 28 19 15 15 25 25 40 40 54 54 32 32 11 13 12 22 22 29 31 25 25 27 25 25 23 23 22 23 25 25 30 21 21 25 25 17 17 16 16 12 12 47 55 73 73 570 570 1323 1323 921 921 370 380 380 985 985 570 570 1439 1439 988 988 167 222 433 204 204 436 485 283 500 280 360 360 561 561 230 240 420 610 635 421 421 210 210 170 170 130 130 201 201 1287 1501 2317 2317 30 30 49 49 32 32 20 10 10 20 20 30 30 47 47 37 37 14 11 11 18 18 28 29 30 25 34 25 30 30 30 35 16 17 26 27 30 30 20 20 12 12 12 12 12 12 49 54 82 82 EN-42 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.5 Table 5: Error messages Error-No.: Description 1 Imbalance arose 2 Imbalance sensor is defective 4 Imbalance switch has been activated for longer than 5 seconds 8 Transponder in the rotor is defective 11 Temperature sensor is defective 12 Chamber over temperature 14 Leap of speed is too big between two mesaurements CLOSE lid 33 34 38 40 Open lid while motor is running Lid contact defective Lid motor is blocked Communication with frequency converter distrubed during start 41 Communication with frequency converter distrubed during stop 42 Short circuit in the frequency converter 43 Undervoltage frequency converter 44 Overvoltage frequency converter 45 Over temperature frequency converter 46 Over temperature motor 47 Over current frequency converter 48 Timeout between control unit and frequency converter 49 Other error frequency converter 55 Overspeed 70 Timeout between controler and RS232 interface 99 Rotor is not allowed in this centrifuge FALSE rotor no Inserted rotor does not exist in the programm Rotor is not detected FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R EN-43 11.6 Table 6: Radius correction and adapter specifications Compatible FC5816R FC5916 FC5916R Rotor, Sw ing out, 4x 250ml, ID Blank NO Bucket/Adapter 8.5 ● ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.3 ● ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 -1.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 16.7 ● ● 30314903 Rack, 1x 250ml D62mm FB, 2/pk 15.7 -1.0 ● ● 30314904 Rack, 1x 100ml D41mm RB, 2/pk 15.9 -0.8 ● ● 30314905 Rack, 3x 50ml D29mm RB, 2/pk 16.3 -0.4 ● ● 30314906 Rack, 3x 50ml D29mm FA, 2/pk 16.7 0.0 ● ● 30314907 Rack, 1x 50ml D34mm RB, 2/pk 15.8 -0.9 ● ● 30314908 Rack, 2x 50ml D29mm Rim, 2/pk 15.8 -0.9 ● ● 30314909 Rack, 1x 50ml D29mm Rim, 2/pk 16.1 -0.6 ● ● 30314910 Rack, 5x 25ml D24.5mm RB, 2/pk 16.1 -0.6 ● ● 30314911 Rack, 3x 25ml D24.5mm Rim, 2/pk 15.5 -1.2 ● ● 30314912 Rack, 9x 15ml D17mm RB, 2/pk 16.3 -0.4 ● ● 30314913 Rack, 7x 15ml D17mm FA, 2/pk 16.4 -0.3 ● ● 30314914 Rack, 7x 15ml D17mm Rim, 2/pk 16.1 -0.6 ● ● 30314915 Rack, 8x 10ml D16mm RB, 2/pk 16.3 -0.4 ● ● 30314916 Rack, 14x 5-7ml D13mm RB, 2/pk 15.9 -0.8 ● ● 30314917 Rack, 10x 1.6-7ml D13mm Vac, 2/pk 15.9 -0.8 ● ● 30304367 Rack, 8x 4-10ml D16mm Vac, 2/pk 16.3 -0.4 ● ● 30304368 Adapter, 9x 1.5/2.0ml D11mm, 2/pk 15.7 -1.0 ● ● Blank NO Bucket/Adapter 14.1 30304369 30314821 Rotor, Angle, 6x 250ml FB, ID Adapter, 8x 1.5ml D11mm, 2/pk ● ● ● ● 13.2 -0.9 ● ● ● ● 30304370 Adapter, 5x 10ml D16mm RB, 2/pk 13.3 ● ● ● ● 30304371 Adapter, 4x 15ml D17mm FA, 2/pk 13.0 -1.1 ● ● ● ● 12.5 -1.6 ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● 30304372 Adapter, 2x 30ml D26mm RB, 2/pk 30304373 Adapter, 1x 50ml D29mm RB, 2/pk 12.0 -2.1 30304374 Adapter, 1x 50ml D30mm FA, 2/pk 11.7 -2.4 30559412 Adapter 15x 1.5/2.0ml D11mm MTBx 2 V1 9.5 13.6 11.8 Blank Adapter 1x 250ml D60mm CBCx 3 V1 Adapter 1x 175/225ml D61mm FAx 2 V1 NO Bucket/Adapter -4.6 -0.5 -0.5 ● ● ● 30314842 Rack, 1x 100ml D46mm RB, 2/pk 14.1 -0.7 ● ● ● 30314843 Rack, 1x 85ml D38mm RB, 2/pk 14.1 -0.7 ● ● ● 30314844 Rack, 1x 50ml D34mm RB, 2/pk 14.2 -0.6 ● ● ● 30314845 Rack, 1x 50ml D29mm RB, 2/pk 14.2 -0.6 ● ● ● 30314846 Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.7 ● ● ● 30314847 Rack, 1x 50ml D29mm Rim, 2/pk 14.3 -0.5 ● ● ● 30314848 Rack, 1x 30ml D25mm Rim, 2/pk 13.8 -1.0 ● ● ● 30314849 Rack, 1x 25ml D25mm RB, 2/pk 14.4 -0.4 ● ● ● 30314850 Rack, 2x 15ml D17mm FA, 2/pk 14.8 0.0 ● ● ● 30314851 Rack, 1x 15ml D17mm Rim, 2/pk 14.5 -0.3 ● ● ● 30559414 30314822 Bucket/Adapter Description -0.8 30559413 Rotor, Sw ing out, 4x 100ml, ID Correction (cm) FC5816 30314820 Radius (cm) FC5718R 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable Bucket/Ad apter FC5718 Rotor Description FC5714 Rotor 12.8 -2.3 14.8 EN-44 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314852 Rack, 7x 15ml D17mm RB, 2/pk 14.5 -0.3 ● ● ● 30314853 Rack, 9x 5-7ml D13mm RB, 2/pk 14.2 -0.6 ● ● ● 30314854 Rack, 10x 1.5ml D11mm, 2/pk 9.9 -4.9 ● ● ● 30314855 Rack, 9x 4.5ml D12mm FB, 2/pk 14.0 -0.8 ● ● ● 30314856 Rack, 7x 5-7ml D13mm RB, 2/pk 11.6 -3.2 ● ● ● 30314857 Rack, 7x 5-7ml D13mm Vac, 2/pk 14.2 -0.6 ● ● ● 30314858 Rack, 5x 4-10ml D16mm Sar, 2/pk 14.3 -0.5 ● ● ● Blank 14.6 ● ● ● 30314864 NO Bucket/Adapter Bucket, 100ml w o Cap, Sealable, 2/pk Bucket, 100ml w / Cap, Sealable, 2/pk Rack, 1x 100ml D40mm RB, 2/pk 30314865 30314866 30314860 14.2 -0.4 ● ● ● 14.2 -0.4 ● ● ● 13.8 -0.8 ● ● ● Rack, 1x 85ml D38mm RB, 2/pk 14.0 -0.6 ● ● ● Rack, 1x 50ml D34mm RB, 2/pk 14.0 -0.6 ● ● ● 30314867 Rack, 1x 50ml D29mm RB, 2/pk 14.0 -0.6 ● ● ● 30314868 Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.5 ● ● ● 30314869 Rack, 1x 50ml D29mm Rim, 2/pk 13.9 -0.7 ● ● ● 30314870 Rack, 1x 30ml D25mm Rim, 2/pk 13.1 -1.5 ● ● ● 30314871 Rack, 1x 25ml D25mm RB, 2/pk 14.0 -0.6 ● ● ● 30314872 Rack, 1x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314873 Rack, 2x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314874 Rack, 2x 15ml D17mm Rim, 2/pk 14.0 -0.6 ● ● ● 30314875 Rack, 4x 15ml D17mm RB, 2/pk 14.0 -0.6 ● ● ● 30314876 Rack, 7x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314877 Rack, 5x 1.5ml D11mm, 2/pk 14.0 -0.6 ● ● ● 30314878 Rack, 3x 10ml D17mm Sar, 2/pk 14.0 -0.6 ● ● ● 30314879 Rack, 4x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314880 14.0 -0.6 ● ● ● 14.6 0.0 ● ● ● 14.6 0.0 ● ● ● 30314883 Rack, 4x 10ml D17mm Vac, 2/pk Bucket, 2x 50ml D29mm FA, w / Rack, 2/pk Bucket, 3x 15ml D17mm FA, w / Rack, 2/pk Cy to, Rectangular Bucket, 2/pk 6.3 -8.3 ● ● ● Blank NO Bucket/Adapter 12.0 ● ● ● ● ● ● ● 30314891 Adapter, 48x 1.5/2.0ml D11mm, 2/pk 12.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 10.3 ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 30314894 30314895 30314861 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314881 30314882 30314824 30314825 30314826 Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID 0.0 9.6 -0.7 ● ● Adapter, 1x 30ml D25mm RB, 2/pk 9.6 -0.7 ● ● Adapter, 1x 50ml D29mm RB, 2/pk 10.0 -0.3 ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 9.8 -0.5 ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 9.3 -1.0 ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 10.3 0.0 ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 9.5 -0.8 ● ● Blank NO Bucket/Adapter 11.5 ● ● ● ● ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 10.9 -0.6 ● ● ● ● ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 11.1 -0.4 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314827 30314828 30314829 30314830 30314831 30314832 30314834 30314835 30314836 30314838 30314839 Rotor, Angle, 4x 85ml RB, ID, Hi Rotor, Sw ing out, 16x 50ml, ID Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable Rotor, Angle, 24x 1.5/2.0ml, ID Rotor, Angle, 12x 1.5/2.0ml, ID Blank NO Bucket/Adapter 9.2 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 8.5 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 30314896 EN-45 ● ● ● ● ● ● -0.7 ● ● ● ● ● ● 8.4 -0.8 ● ● ● ● ● ● 8.9 -0.3 ● ● ● ● ● ● Adapter, 1x 50ml D29mm FA, 2/pk 8.6 -0.6 ● ● ● ● ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 8.3 -0.9 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 7.5 -1.7 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 8.3 -0.9 ● ● ● ● ● ● Blank NO Bucket/Adapter 16.5 ● ● 30304375 Rack, 4x 50ml D29mm FA, 2/pk 16.5 0.0 ● ● 30314583 Rack, 10x 15ml D17mm FA, 2/pk 16.5 0.0 ● ● Blank NO Bucket/Adapter 13.0 30130894 Adapter, 1x 15ml D17mm RB, 2/pk 12.8 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 12.2 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk Blank NO Bucket/Adapter 11.0 30130891 Adapter, 1x 30ml D26mm, 2/pk 10.7 30130892 Adapter, 1x 16ml D18mm, 2/pk 10.3 30130893 Adapter, 1x 15ml D17mm, 2/pk 30130894 30130889 30130890 30130886 ● ● ● ● ● ● -0.2 ● ● ● ● ● ● -0.8 ● ● ● ● ● ● 10.4 -2.6 ● ● ● ● ● ● 8.9 -4.1 ● ● ● ● ● ● ● ● ● -0.3 ● ● ● -0.7 ● ● ● 10.6 -0.4 ● ● ● Adapter, 1x 15ml D17mm RB, 2/pk 10.6 -0.4 ● ● ● Adapter, 1x 7ml D13.5mm RB, 2/pk 10.2 -0.8 ● ● ● Adapter, 1x 5ml D13.5mm, 2/pk 8.3 -2.7 ● ● ● Adapter, 1x 1.5/2.0ml D11mm, 6/pk 6.7 -4.3 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130891 Adapter, 1x 30ml D26mm, 2/pk 8.2 -0.2 ● ● ● ● ● ● 30130892 Adapter, 1x 16ml D18mm, 2/pk 7.9 -0.5 ● ● ● ● ● ● 30130893 Adapter, 1x 15ml D17mm, 2/pk 8.0 -0.4 ● ● ● ● ● ● 30314892 Adapter, 1x 15ml D17.5mm FA, 2/pk 7.7 -0.7 ● ● ● ● ● ● Blank NO Bucket/Adapter 12.5 ● ● ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 12.2 -0.3 ● ● ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 10.5 -2.0 ● ● ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 9.0 -3.5 ● ● ● ● ● Blank NO Bucket/Adapter 11.0 ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 10.6 -0.4 ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 9.1 -1.9 ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 7.7 -3.4 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.2 ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.7 -0.7 ● ● ● ● ● ● Blank NO Bucket/Adapter 9.4 ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 9.1 -0.3 ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 8.4 -1.0 ● ● ● ● ● Blank NO Bucket/Adapter 8.5 ● ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.3 ● ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 -1.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 6.5 ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 6.3 -0.2 ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 5.6 -0.9 ● ● EN-46 FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Blank Adapter, 1x 1.5ml for ov er16000x g, 6/pk NO Bucket/Adapter 19.3 30553138 Adapter 32x 7ml D13mm RBx 2 V1 18.7 30553135 Adapter 18x 15ml D17mm RBx 2 V1 30553132 Adapter 8x 30ml D26mm RBx 2 V1 30553131 30314900 30553085 Rotor Sw ingOut 4x 750ml D99mm ID SL ● ● ● -0.6 ● ● 18.8 -0.5 ● ● 19.1 -0.2 ● ● Adapter 7x 50ml D29mm RBx 2 V1 19.1 -0.2 ● ● 30553127 Adapter 3x 100ml D41mm RBx 2 V1 19.1 -0.2 ● ● 30553125 18.9 -0.4 ● ● 19.1 -0.2 ● ● 30553128 Adapter 1x 250ml D62mm FBx 2 V1 Adapter 1x 225/190ml D61mm FAx 2 V1 Adapter 5x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553130 Adapter 7x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553133 Adapter 14x 15ml D17mm FAx 2 V1 19.2 -0.1 ● ● 30553134 Adapter 17x 15ml D17mm FAx 2 V1 Adapter 18x 5/10ml D16mm Vac/Sarx 2 V1 Adapter 38x 1.1/1.4ml D8mm Sarx 2 V1 19.0 -0.3 ● ● 19.0 -0.3 ● ● 18.9 -0.4 ● ● 18.9 -0.4 ● ● 18.8 -0.5 ● ● ● ● 30559421 Adapter 21x 10ml D16mm RBx 2 V1 Adapter 27x 1.8/7ml D13mm Vac/Sarx 2 V1 NO Bucket/Adapter Adapter 1x 250ml D61.5mm FBx 2 V1 Adapter 1x 175/225ml D61.5mm FAx 2 V1 Adapter 1x 50ml D29.5mm FAx 2 V1 Adapter 1x 250ml D60.5mm CBCx 2 V1 Adapter 4x 30ml D25.5mm RBx 2 V1 30559420 Adapter 3x 50ml D29mm RBx 2 V1 30559422 Adapter 6x 15ml D17mm FAx 2 V1 30553126 30559377 30553136 30553139 Blank 30564850 30559417 30553086 -0.1 ● 30553140 Rotor Angle 4x 500ml D69mm ID 6.4 30559419 30559416 14.5 13.7 -0.8 ● ● 13.4 -1.1 ● ● 12.4 -2.1 ● ● 12.6 -1.9 ● ● 13.8 -0.7 ● ● 14.3 -0.2 ● ● 13.8 -0.7 ● ● EN-47 11.7 Table 8: Redemption form / Decontamination certificate Enclose this form with all returns of equipment and assemblies! The completed declaration about the decontamination is a prerequisite for the assumption and further processing of the return. If no corresponding explanation is enclosed, we carry out decontamination with costs at your expense. Please fill out in block capitals!  FrontierTM Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Surname; last name: ___________________________________________ Organization / company: ________________________________________ Street: _____________________________________ _________________ ZIP CODE: _____________ place:____________________ Telephone: _____________ fax:____________________ E-Mail: _____________________________________ Pos. Crowd Decontaminated object Serial number Description / Comment 1 2 3 4 Are the parts listed above in contact with the following substances? Health endangering watery solutions, buffers, acids, alkalis:…………………….....  Yes  No Potentially infectious agents: ……………………………………………………………  Yes  No Organic reagents and solvent: ……………………………………………………….. ..  Yes  No Radioactive substances: ……………………………………………. α.. β.. γ…  Yes  No Health endangering proteins: …………………………………………………………..  Yes  No DNA: ………………………………………………………………………………………  Yes  No These substances have reached the equipment/assembly? ……………………….  Yes  No Which one, if yes: ________________________________________________________________ Description of the measures for the decontamination of the listed parts: I confirm the proper decontamination: Company/Dept .________________________ Place and Date: ____________________________________ Signature of the authorized person:________________________________________ Croquis del FC5714 / FC5718 / FC5816 / FC5916 1 2 7 4 3 6 5 Figura 1 Croquis del FC5718R 1 2 7 3 4 6 5 Figura 2 Croquis del FC5816R / FC5916R 1 2 7 3 6 4 5 Figura 3 1 Tapa de la centrífuga 3 Pantalla 5 Interruptor de corriente principal 7 Liberación de emergencia 2 Ventana del rotor 4 Etiqueta de funciones 6 Conexión de energía Etiqueta de funciones Etiqueta de funciones para FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figura 4 1 4 5 6 7 8 2 3 9 10 11 12 Figura 5 1. 3. 5. 7. Pantalla LCD Inicio de centrifugación Selección de intensidad Aceleración/Desaceleración Configuración de tiempo 9. Centrifugación rápida 11. Detener centrifugación / configuración 2. 4. 6. Configuración de programa Selección de RPM/RCF Configuración de temperatura (Solo FC5515R) 8. Guardar información de configuración. 10. Abrir puerta 12. Perilla de ajuste Pantalla LCD La siguiente imagen muestra los elementos individuales de la pantalla LCD. Figura 6 M2 M3 M1 M12 M15 M13 M4 M9 M11 M10 M5 M6 M7 A-1 M8 A-2 M14 A-3 A-4 Figura 7 Campos de la pantalla: A-1 A-2 A-3 A-4 Campo de pantalla – "RPM/RCF" Campo de pantalla – "Acc/Dec" "Service" Campo de pantalla – "Time/Prog" Campo de pantalla –"Temp" Mensajes/logotipos de los campos de la pantalla M1 M4 "Cerrado" "Número de rotor" M2 M5 "Abierto" "Revoluciones por Minuto" M3 M6 M7 M10 M13 "Aceleración" "Programa" "h m s" M8 M11 M14 "Desaceleración" "Error" "temperatura" M9 M12 M15 "Rotor" "Fuerza de centrifugación relativa" "Radio" "Servicio" "preenfriamiento" Tabla de números de rotor FC5816R FC5916 FC5916R Rotor, Angle, 24x1.5/2.0ml, ID, Sealable Rotor, Swing out, 4x250ml, ID Rotor, Angle, 6x250ml FB, ID Rotor, Swing out, 4x100ml, ID Rotor, Swing out, 4x100ml, ID, Sealable Rotor, Swing out, 2x3MTP, ID Rotor, Angle, 6x85ml RB, ID, Hi Rotor, Angle, 6x85ml RB, ID Rotor, Angle, 4x85ml RB, ID, Hi Rotor, Swing out, 16x50ml, ID Rotor, Angle, 10x50ml FA, ID Rotor, Angle, 6x50ml RB/FA, ID Rotor, Angle, 6x50ml RB, ID, Hi Rotor, Angle, 30x15ml RB/FA, ID Rotor, Angle, 20x10ml RB, ID, Hi Rotor, Angle, 12x15ml RB/FA, ID Rotor, Angle, 44x1.5/2.0ml, ID Rotor, Angle, 30x1.5/2.0ml, ID, Sealable Rotor, Angle, 24x1.5/2.0ml, ID Rotor, Angle, 12x1.5/2.0ml, ID Rotor, Angle, 64x0.5ml, ID Rotor, Angle, 4x8-w PCR Strip, ID Rotor SwingOut 4x750ml D99mm ID SL Rotor Angle 4x500ml D69mm ID FC5816 30304361 30314820 30314821 30314822 30314823 30314824 30314825 30314826 30314827 30314828 30314829 30314830 30314831 30314832 30314833 30314834 30314835 30314836 30314838 30314839 30314840 30314841 30553085 30553086 FC5718R 61 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 38 39 40 41 85 86 Capacidad FC5718 N.° de orden Compatibilidad FC5714 Indicador del n.° del rotor ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● 1. 2. 3. 4. 5. TABLA DE CONTENIDO INTRODUCCIÓ N ........................................................................................................................................................................... 1 1.1 DESCRIPCIÓ N ................................................................................................................................................................................... 1 1.2 CARACTERÍSTICAS ............................................................................................................................................................................. 1 1.3 DEFINICIÓ N DE LETREROS Y SÍMBOLOS DE ADVERTENCIA......................................................................................................................... 1 1.4 PRECAUCIONES DE SEGURIDAD ........................................................................................................................................................... 2 1.4.1 Usuario................................................................................................................................................................................ 2 1.4.2 Rotor y accesorios.............................................................................................................................................................. 2 1.4.3 Medidas para su protección.............................................................................................................................................. 2 1.4.4 Excluya las siguientes influencias ambientales .............................................................................................................. 3 1.4.5 Medidas de seguridad operativa...................................................................................................................................... 3 1.4.6 Peligro y precauciones....................................................................................................................................................... 3 INSTALACIÓ N ............................................................................................................................................................................... 4 2.1 DESEMPACADO ................................................................................................................................................................................ 4 2.1.1 Paquete de entrega ........................................................................................................................................................... 4 2.2 SELECCIÓ N DE LA UBICACIÓ N .............................................................................................................................................................. 4 2.3 I NSTALACIÓ N ................................................................................................................................................................................... 4 2.4 PRECAUCIONES DE SEGURIDAD DURANTE LA OPERA CIÓ N Y LA GARA NTÍA .................................................................................................. 5 OPERACIÓ N .................................................................................................................................................................................. 5 3.1 MONTAJE Y CARGA DEL ROTOR ........................................................................................................................................................... 5 3.1.1 Instalación de rotores........................................................................................................................................................ 5 3.1.2 Carga de los rotores angulares......................................................................................................................................... 6 3.1.2 Carga de los rotores oscilantes......................................................................................................................................... 6 3.1.4 Carga y sobrecarga de rotores.......................................................................................................................................... 7 3.1.5 Remoción del rotor ............................................................................................................................................................ 7 3.2 CONTROL DE LA TAPA ........................................................................................................................................................................ 7 3.2.1 Tapa abierta ....................................................................................................................................................................... 7 3.2.2 Seguro de la tapa ............................................................................................................................................................... 8 3.3 PRESELECCIÓ N ................................................................................................................................................................................. 8 3.3.1 Preselección de velocidad / valor RCF.............................................................................................................................. 8 3.3.2 Preselección del tiempo de operación ............................................................................................................................. 9 3.3.3 Preselección de intensidad del freno y aceleración........................................................................................................ 9 3.3.4 Preselección de temperatura (solo para modelos refrigerados)................................................................................. 10 3.3.5 Preenfriamiento (solo para modelos refrigerados)...................................................................................................... 10 3.4 CORRECCIÓ N DEL RADIO .................................................................................................................................................................. 10 3.5 PROGRAMA ................................................................................................................................................................................... 11 3.5.1 Almacenamiento de programas..................................................................................................................................... 11 3.5.2 Recuperación de programas almacenados ................................................................................................................... 12 3.5.3 Salir del modo de programa ........................................................................................................................................... 12 3.6 ARRANQUE Y PARO DE LA CENT RÍFUGA .............................................................................................................................................. 12 3.6.1 Arranque de la centrífuga ............................................................................................................................................... 12 3.6.2 Botón "STOP".................................................................................................................................................................... 13 3.7 DETECCIÓ N DE DESBALANCE ............................................................................................................................................................. 13 CONFIGURACIÓ N ....................................................................................................................................................................... 13 4.1 AJUSTES BÁSICOS............................................................................................................................................................................ 13 4.1.1 Acceso al modo "Operating Data" ................................................................................................................................. 13 4.1.2 Indicación de temperatura.............................................................................................................................................. 14 4.1.3 Encendido y apagado de la señal................................................................................................................................... 15 4.1.4 Preselección del volumen de la señal de sonido ........................................................................................................... 15 4.1.5 Selección de melodía para la señal del sonido, final de operación ............................................................................ 15 4.1.6 Encendido y apagado del sonido del teclado ................................................................................................................ 16 4.1.7 Extracción de datos operativos ...................................................................................................................................... 16 MANTENIMIENTO...................................................................................................................................................................... 17 5.1 MANTENIMIENTO Y LIMPIEZA........................................................................................................................................................... 17 5.1.1 Cuidado general: ............................................................................................................................................................. 17 5.1.2 Limpieza y desinfección de la unidad............................................................................................................................. 18 5.1.3 Limpieza y desinfección del rotor ................................................................................................................................... 18 5.1.4 Desinfección de rotores de aluminio.............................................................................................................................. 18 5.1.5 Desinfección de rotores de PP......................................................................................................................................... 18 5.1.6 Ruptura de vidrio ............................................................................................................................................................. 19 5.2 VIDA ÚTIL DE ROTORES , PORTATUBOS Y ACCESORIOS ........................................................................................................................... 19 6. IDENTIFICACIÓ N Y SOLUCIÓ N DE PROBLEMAS ...................................................................................................................... 19 6.1 MENSAJE DE ERROR: CAUSA / SOLUCIÓ N........................................................................................................................................... 19 6.2 ESTUDIO DE POSIBLES MENSAJES DE ERROR Y SUS SOLUCIONES .............................................................................................................. 19 6.2.1 Apertura de la tapa durante una falla de la corriente eléctrica (Apertura de emergencia de la tapa) ................. 19 6.2.2 Descripción del sistema de mensajes de error .............................................................................................................. 20 7. RECEPCIÓ N DE CENTRÍFUGAS PARA REPARACIÓ N ................................................................................................................ 21 8. TRANSPORTE, ALMACENAMIENTO Y ELIMINACIÓ N.............................................................................................................. 21 8.1 TRANSPORTE.................................................................................................................................................................................. 21 8.2 ALMACENAMIENTO......................................................................................................................................................................... 21 9. FICHA TÉCNICA........................................................................................................................................................................... 23 9.1 ESPECIFICACIONES .......................................................................................................................................................................... 23 9.1.1 Centrifuge FC5714 ............................................................................................................................................................ 23 9.1.2 Centrífuga FC5718............................................................................................................................................................ 24 9.1.3 Centrifuge FC5718R.......................................................................................................................................................... 25 9.1.6 Centrifuge FC5916 ............................................................................................................................................................ 28 9.1.7 Centrifuge FC5916R.......................................................................................................................................................... 29 9.2 DIAGRAMAS Y DIMENSIONES ................................................................................................................................................... 30 10. CONFORMIDAD..................................................................................................................................................................... 31 11. APÉNDICE............................................................................................................................................................................... 32 11.1 TABLE 1: PESO NETO PERMITIDO ............................................................................................................................................. 32 11.2 TABLE 2: TEMPERATURAS MÍNIMAS A MÁXIMA VELOCIDAD ͘͘͘͘.........................................................................................͘͘34 11.3 TABLE 3: MÁXIMA VELOCIDAD Y VALORES RCF PARA ROTORES PERMITIDOS............................................................͘͘͘͘͘͘͘ ͘͘35 11.4 TABLE 4: TIEMPOS DE ACELERACIÓ N Y DESACELERACIÓ N..................................................................................................... 38 11.5 TABLA 5: MENSAJES DE ERROR....................................................................................................................................................... 40 11.6 TABLE 6: CORRECCIÓ N DEL RADIO ........................................................................................................................................... 41 11.7 TABLA 7: FORMULARIO DE DEVOLUCIÓ N / CERTIFICADO DE DESCONTAMINACIÓ N ................................................................................. 45 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-1 1. INTRODUCCIÓN 1.1 Descripción Gracias por elegir este producto OHAUS. Todos los símbolos indican instrucciones de seguridad y puntos de situaciones peligrosas potenciales. Lea el manual por completo antes de usar las centrífugas Frontier TM MultiPro para evitar un funcionamiento incorrecto. Las centrífugas FrontierTM MultiPro fueron diseñadas para la separación de materiales o mezclas con diferentes densidades. Las centrífugas OHAUS están diseñadas exclusivamente para funcionar en interiores y para ser utilizadas solo por personal cualificado. 1.2 Características El modelo FC5714/FC5718/FC5816/FC5916 es una centrífuga universal no refrigerada. El modelo FC5718R/FC5816R/FC5916R es una centrífuga universal refrigerada. Los tres modelos que ofrecemos pueden funcionar en dos voltajes, 230 V o 120 V. Las centrífugas pueden utilizarse con rotores angulares y basculantes. Todos los parámetros son accesibles a través de botones y se seleccionan con el ajustador central. Todos los valores preseleccionados y actuales se muestran permanentemente en la pantalla LCD. La centrífuga es impulsada por un motor de inducción libre de mantenimiento. Los datos técnicos detallados se pueden encontrar en la sección "Ficha técnica". 1.3 Definición de letreros y símbolos de advertencia Las notas de seguridad están marcadas con palabras de señalamiento y símbolos de advertencia. Estos muestran problemas y advertencias de seguridad. Ignorar las notas de seguridad puede resultar en lesiones, daño al instrumento, mal funcionamiento y resultados falsos. El grado de peligro es parte de una nota de seguridad y distingue los posibles resultados de la falta de observancia. Palabras de señalamiento PELIGRO Conducirá a lesiones graves o muerte si no se evita. ADVERTENCIA Para una situación peligrosa con riesgo medio, que posiblemente resulta en lesiones o muerte si no se evita. PRECAUCIÓN Para una situación peligrosa con riesgo bajo, que resulta en daño al dispositivo, a la propiedad o en pérdida de datos si no se evita. ATENCIÓN Para información importante acerca del producto. Puede resultar en daño del equipo si no se evita. NOTA Para información útil acerca del producto. Símbolos de advertencia Riesgo general Corriente alterna Explosión Riesgo de descarga eléctrica Riesgo biológico Aplastamiento ES-2 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Letreros de advertencia e información en la superficie de la centrífuga Se deben usar cuatro portatubos siempre en los rotores basculantes de cuatro lugares, o de lo contrario ocurrirá daño a la centrífuga. Este daño no estará cubierto por la garantía del producto. ¡Atención! Revise el apriete de la tuerca del rotor antes de cada uso. Desconecte el enchufe de la corriente principal antes de abrir la carcasa o la apertura de emergencia. Dirección de rotación – el impulsor del rotor tiene rotación a la derecha Referencia para carga de rotores 1.4 Precauciones de seguridad 1.4.1 Usuario Las centrífugas OHAUS están diseñadas exclusivamente para uso en interiores y para uso por parte de personal calificado. Este dispositivo solo puede ser usado por personal especialista capacitado. Dicho personal debe haber leído el manual de operación y estar familiarizado con el funcionamiento del dispositivo. 1.4.2 Rotor y accesorios Solo deben usarse rotores y accesorios OHAUS originales. Cualquier otro uso o uso intencional se considera inadecuado. OHAUS no se hace responsable por daños que resulten del uso inadecuado. PRECAUCIÓN: Lea todas las advertencias de seguridad antes de instalar, hacer conexiones o dar servicio a este equipo. La falta de cumplimiento de estas advertencias podría resultar en lesiones o daños a la propiedad. Conserve las instrucciones para referencia futura. 1.4.3 Medidas para su protección ADVERTENCIA: ¡Nunca trabaje en un entorno sujeto a riesgos de explosión! La caja del instrumento no es hermética al gas. (Riesgo de explosión debido a la generación de chispas y corrosión causada por el ingreso de gases.) ADVERTENCIA: Cuando use químicos y solventes, siga las instrucciones del fabricante y las reglas generales de seguridad de laboratorio. ADVERTENCIA: La centrífuga no está sellada. Adopte medidas de protección adecuadas cuando use la centrífuga para muestras infecciosas o patógenas. Tome precauciones de seguridad adecuadas cuando maneje muestras de estos tipos. Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-3 1.4.4 Excluya las siguientes influencias ambientales       Vibraciones fuertes Luz solar directa Humedad atmosférica mayor del 80% Presencia de gases corrosivos Temperaturas menores de 2 °C y mayores de 35 °C Campos eléctricos o magnéticos intensos: ADVERTENCIA: Existe el riesgo de descarga eléctrica dentro de la caja. Solo personal autorizado y calificado debe abrir la caja. Desconecte todas las conexiones eléctricas hacia la unidad antes de abrirla. 1.4.5 Medidas de seguridad operativa  No destornille las dos mitades de la caja.  ¡Seque cualquier derrame de líquidos inmediatamente! El instrumento no es hermético al agua.  Verifique que el rango de voltaje de entrada y tipo de enchufe sean compatibles con la alimentación eléctrica local.  Conecte el cable de corriente únicamente a un enchufe adecuado con toma de tierra.  Utilice únicamente un cable de corriente con una potencia que supere las especificaciones del etiquetado del equipo.  No coloque el equipo de forma que resulte difícil desconectar el cable de corriente del enchufe.  Verifique que el cordón eléctrico no represente un obstáculo potencial o riesgo de tropezarse.  El equipo es solo para uso en interiores. Use el equipo solamente en lugares secos.  Use solo accesorios aprobados.  Opere el equipo sólo en las condiciones ambientales especificadas en estas instrucciones  Desconecte el equipo de la corriente eléctrica cuando lo limpie.  No opere el equipo en ambientes peligrosos o inestables.  El servicio debe proporcionarse solamente por personal autorizado. 1.4.6 Peligro y precauciones Para proteger a las personas y el entorno, siga estas precauciones:  Durante la centrifugación, se prohíbe la presencia de personas y la manipulación de materiales peligrosos dentro de 30 cm alrededor de la centrífuga, de acuerdo con las regulaciones de EN 61010-2-020.  La FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R no es a prueba de explosión y por lo tanto no debe operarse en áreas o lugares con riesgo de explosión. La centrifugación de sustancias inflamables, explosivas, radioactivas o similares, que reaccionan en forma química con la energía, está estrictamente prohibida. La decisión final sobre los riesgos asociados con el uso de tales sustancias es responsabilidad del usuario de la centrífuga.  Nunca centrifugue material tóxico o patógeno sin las precauciones de seguridad adecuadas; por ejemplo, la centrifugación de portatubos o tubos cuyo sello hermético no está presente o es defectuoso está estrictamente prohibida. El usuario está obligado a implementar procedimientos de desinfección adecuados en caso de que sustancias peligrosas contaminen la centrífuga o sus accesorios. Cuando centrifugue sustancias infecciosas, siempre preste atención a las precauciones generales de laboratorio. ¡Si es necesario, póngase en contacto con su representante de seguridad!  Se prohíbe operar la centrífuga con rotores que no sean los indicados para esta unidad.  Bajo ninguna circunstancia abra la tapa de la centrífuga mientras el rotor esté aún funcionando o girando a una velocidad > 2 m/s. ES-4 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 1.4.7 Abreviaturas usadas en este manual Símbolo/Abrev iaturas RPM RCF PCR PP PC accel decel Prog Unidad [min -1] rpm [x g] - Descripción revoluciones por minuto fuerza centrífuga relativa PCR Reacción en cadena de la polimerasa Polipropileno Policarbonato aceleración desaceleración programa 2. INSTALACIÓN 2.1 Desempacado Retire con cuidado la centrífuga y cada uno de sus componentes del paquete. Los componentes incluidos varían de acuerdo con el modelo de centrífuga (vea la tabla siguiente). Guarde el empaque para poder almacenar y transportar la centrífuga adecuadamente. ¡El manual de instrucciones debe mantenerse siempre con la centrífuga! Los rotores y accesorios se empaquetan aparte. ADVERTENCIA: Peligro de levantamiento. Existe riesgo de lesiones si una sola persona la levanta. Solicite ayuda al levantar o mover el equipo. Consulte la sección 8.3 para obtener detalles sobre cómo sacar el equipo del embalaje. Se recomienda que cuatro o más personas transporten e instalen estas tres unidades, o utilicen algún equipo de transporte. Consulte la sección 8.3 para ver más detalles sobre cómo levantar estas unidades cuando están con embalaje. 2.1.1 Paquete de entrega Cantidad 1 Descripción Centrífuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Cordón eléctrico Tarjeta de garantía Manual de instrucciones / Guía rápida Llave del rotor 1 1 1 1 2.2 Selección de la ubicación ¡NOTA! Evite las vibraciones excesivas, fuentes de calor, corrientes de aire y cambios rápidos de temperatura.  La centrífuga debe instalarse en una superficie sólida, nivelada y uniforme, si es posible en un gabinete o mesa de laboratorio, o en alguna otra superficie libre de vibraciones.  Durante la centrifugación, la centrífuga debe colocarse de tal forma que quede un espacio mínimo de 30 cm en cada lado de la unidad, de acuerdo con los estándares EN 61010-2-020.  No coloque la centrífuga cerca de una ventana o calentador donde pudiera estar expuesta al calor excesivo, ya que el funcionamiento de esta se basa en una temperatura ambiente de 23 °C. 2.3 Instalación Siga estos pasos:  Verifique si la corriente eléctrica corresponde a la especificada en la etiqueta de clasificación del fabricante, la cual está en el panel posterior.  Para el modelo FC5714/FC5718/FC5816, la línea de alimentación debe estar protegida por un disyuntor de 10 A circuito de calificación (tipo K).  Para los modelos FC5718R/FC5816R /FC5916/FC5916R, la línea de alimentación debe estar protegida por un disyuntor Una calificación de 16 (tipo K).  Para caso de emergencia, debe haber un interruptor de emergencia instalado fuera de la sala para desconectar la corriente eléctrica de la unidad.  Conecte la centrífuga en la toma de corriente eléctrica con conexión a tierra.  Conecte la centrífuga al suministro eléctrico. Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-5 (La toma para el cable de alimentación debe ser fácil de alcanzar y también fácil de desconectar)  Encienda el instrumento usando el interruptor de corriente eléctrica.  Abra la tapa con el botón "Door Open".  Retire el dispositivo de fijación de transporte del motor. 2.4 Precauciones de seguridad durante la operación y la garantía  No opere la centrífuga en caso de que no esté instalada correctamente.  Durante su funcionamiento, no se apoye en la centrífuga.  Por razones operativas, no permanezca más de lo necesario en el radio de 30 cm alrededor del equipo.  No coloque ningún material potencialmente peligroso en el radio de 30 cm alrededor del equipo.  No opere la centrífuga cuando esté desarmada (por ejemplo, sin la caja).  No opere la centrífuga cuando se hayan alterado ilegalmente componentes mecánicos o eléctricos.  No use accesorios como rotores y portatubos que no estén aprobados exclusivamente por OHAUS Corporation, excepto tubos para centrífuga disponibles comercialmente hechos de vidrio o plástico.  No centrifugue sustancias sumamente corrosivas, ya que pueden dañar o debilitar los materiales.  No opere la centrífuga con rotores o portatubos que muestren signos de corrosión o daño mecánico.  El fabricante es responsable de la seguridad y confiabilidad de la centrífuga solamente si:  La unidad se opera de acuerdo con este manual de instrucciones.  Las modificaciones, reparaciones u otros ajustes son realizados por personal autorizado y la instalación eléctrica cumple con el código eléctrico correspondiente. ATENCIÓN: Garantía La centrífuga ha sido sometida a rigurosas pruebas y controles de calidad. En el improbable caso de que ocurra una falla por defecto de fabricación, la centrífuga y los rotores están cubiertos por una garantía. Esta garantía queda sin validez en caso de mal manejo, daños o negligencia, y además en el caso de utilizar piezas de repuesto y/o accesorios incorrectos o realizar modificaciones no autorizadas en la unidad. El fabricante se reserva el derecho de realizar modificaciones técnicas para mejorar el funcionamiento de la unidad. 3. OPERACIÓN 3.1 Montaje y carga del rotor 3.1.1 Instalación de rotores Limpie el eje motriz y el collar con una tela limpia y libre de grasa. Coloque el rotor en el eje motriz. (Vea la figura siguiente). Tenga cuidado de que el rotor esté completamente instalado en el eje del motor. Eje del motor y cámara Figura 8 ES-6 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Tuerca para rotor Figura 9 Herramienta para rotor con tuerca Herramienta para rotor sin tuerca Tapa insertable Tapa atornillable Figura 10 Sostenga el rotor con una mano y fíjelo en el eje al girar la tuerca de fijación en sentido del reloj. Apriete la tuerca de fijación con la llave del rotor incluida (vea las figuras 9-10) Proporcionaremos una herramienta para rotor sin tuerca con la centrífuga; la herramienta para rotor con tuerca se proporcionará con el rotor. ATENCIÓN: Verifique que el tornillo de fijación esté instalado correctamente antes de cada operación. (Vea la figura 11.) No opere la centrífuga con rotores o portatubos que muestren signos de corrosión o daño mecánico. No la opere con sustancias sumamente corrosivas que pudieran dañar el rotor, los portatubos y los materiales. En caso de alguna duda, póngase en contacto con el fabricante. 3.1.2 Carga de los rotores angulares Los rotores deben cargarse en forma simétrica y con pesos iguales (vea la figura siguiente). El adaptador solo puede cargarse con los portatubos adecuados. Las diferencias en peso entre los portatubos dañados deben mantenerse lo más pequeñas que sea posible. Por lo tanto, se recomienda pesarlos con una balanza. Esto reduce el desgaste del motor y el ruido de operación. La carga máxima por orificio está establecida para cada rotor. Figura 11 INCORRECTO Figura 12 CORRECTO (6 tubos) 3.1.2 Carga de los rotores oscilantes La carga de los portatubos debe hacerse de acuerdo a la figura a continuación. Está permitido operar por ejemplo un rotor de 4 lugares con 2 portatubos cargados solamente, pero los portatubos cargados deben estar opuestos entre ellos. Asegúrese de que los portatubos no cargados también se coloquen dentro del rotor (ver la figura abajo). En principio, los rotores basculantes no se pueden hacer funcionar hasta que todos los portatubos o racks se coloquen en el rotor. Los pernos en el rotor deben lubricarse con aceite "High TEF". Los tubos de muestra deben llenarse todos hasta el mismo nivel y deben colocarse en los huecos o racks de tubos. La diferencia de peso de los portatubos cargados no debe exceder aproximadamente 1.0 g. Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-7 ATENCIÓN Los rotores basculantes solo pueden hacerse funcionar si todas las posiciones están llenas con cuatro portatubos o cuatro portavasos – ¡¡no mezcle portatubos y portavasos!! ATENCIÓN No opere la centrífuga con rotores o portatubos que muestren cualquier signo de corrosión o daño mecánico. No opere la unidad con sustancias extremadamente corrosivas, que podrían dañar el rotor y los portatubos. En caso de cualquier pregunta, póngase en contacto con el fabricante. Figure.13 WRONG Figure.14 CORRECT 3.1.4 Carga y sobrecarga de rotores Todos los rotores aprobados con su máxima velocidad y peso de llenado máximo están indicados en la "Tabla 2, peso neto permitido" (vea el APÉNDICE). La carga máxima permitida para un rotor, determinada por el fabricante, así como la máxima velocidad permitida para el rotor (vea la etiqueta en el rotor), no deben excederse. Los líquidos con los que se cargan los rotores deben tener una densidad homogénea máxima de 1.2 g/ml o menos cuando el rotor opera a máxima velocidad. Para centrifugar líquidos de mayor densidad, la velocidad debe reducirse de acuerdo con la siguiente fórmula: Vel. reducida nred = √ 1,2 𝑑𝑒𝑛𝑠𝑖𝑑𝑎𝑑 𝑚𝑎𝑦𝑜𝑟 x máx. vel. (nmáx) del rotor Ejemplo: nred = √ 1,2 1,7 x 4.000 = 3.360 rpm En caso de alguna duda, póngase en contacto con el fabricante. Para determinar la fuerza centrífuga relativa (RCF/fuerza G) para un adaptador específico, puede calcular aplicando DIN 58 970 usando la fórmula siguiente: RCF = 1.117862 *10-5*n2*rmax n: revoluciones por minuto (RPM) rmax: radio de centrifugado máximo en cm usando la parte inferior de los tubos 3.1.5 Remoción del rotor Afloje completamente la tuerca de fijación (tornillo sobre el punto rígido) y levante el rotor en forma vertical fuera de la centrífuga. (Vea las figuras 9 y 10). 3.2 Control de la tapa 3.2.1 Tapa abierta Después de la operación, cuando la tapa de la centrífuga está cerrada, aparece la palabra "close" (M1) en la pantalla "RPM | RCF" (A-1). Además, si hay un rotor en la centrífuga, aparece la palabra "rotor"(M3) así como el número de código del rotor respectivo, el cual está en el sistema de la centrífuga "71" (M4). Si no hay rotor en la centrífuga, destella la palabra "rotor" (M3) y aparece adicionalmente la palabra "no" (M4). Al presionar el botón "Door Open" (7), puede liberar la tapa de la centrífuga. Tan pronto la tapa electromagnética esté completamente liberada, aparece la palabra "open" (M2). Ahora puede abrir la tapa de la centrífuga. Consulte la Figura 15 a continuación para referencia. ES-8 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 1 7 Figura 15 Durante la operación, puede consultar el tipo de rotor en cualquier momento al presionar el botón "Door Open" (7). 3.2.2 Seguro de la tapa La tapa debe bajarse solo un poco. Un seguro de la tapa electromagnétic a cierra la tapa, al mismo tiempo que desaparece la palabra "open" (M2) (vea la Figura 15). Como signo de que la centrífuga está lista para arrancar, en la pantalla "RPM | RCF" (A-1) aparece la palabra "close" (M1). Al mismo tiempo, aparece la palabra "rotor" (M3), así como el número de código del rotor, el cual está en el sistema de la centrífuga, "no 71" (M4). Con eso, todos los datos específicos del rotor, por ejemplo máxima velocidad, aceleración, etc. están adoptados. ATENCIÓN: ¡No deje los dedos entre la tapa y el dispositivo o mecanismo de cierre cuando cierre la tapa! 3.3 Preselección 3.3.1 Preselección de velocidad / valor RCF Esta preselección se activa con el botón "RPM | RCF" (4) (consulte la Figura 16 a continuación). Al presionar el botón una vez, destella la palabra "rpm" (M5). Al presionar el botón dos veces, se puede seleccionar la preselección de las fuerzas centrífugas. Entonces aparece la palabra "rcf" (M6). Puede establecer los valores que desee con la perilla de ajuste (1). En la pantalla (A-1) aparece permanentemente el valor regulado, antes, durante y después de la operación. Figura 16 Mientras no haya rotor insertado, la velocidad es ajustable entre 200 rpm y la máxima revolución de la centrífuga. Si hay un rotor en la centrífuga, la velocidad solo puede preseleccionarse hasta la máxima revolución permitida de ese rotor. Esta es la misma que la preselección del valor RCF. El rango de ajuste es entre la fuerza centrífuga a 200 rpm y la fuerza centrífuga máxima permitida del rotor. Vea la "Tabla 4: máxima velocidad y valores RCF para rotor permitido" (Vea el APÉNDICE). Todos los valores importantes se indican aquí. Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-9 ATENCIÓN: También verifique las revoluciones máximas permitidas de sus tubos de prueba con el fabricante. 3.3.2 Preselección del tiempo de operación El tiempo de operación puede preseleccionarse en tres rangos diferentes desde 10 segundos hasta 99 horas 59 minutos. 1. Rango desde 10 segundos hasta 59 minutos 50 segundos en pasos de 10 segundos 2. Rango desde 1 hora hasta 99 horas 59 minutos en pasos de 1 minuto 3. La operación continua "cont", que puede interrumpirse con el botón "Stop"(10) (vea la Figura 17). El tiempo de operación puede preseleccionarse con la tapa abierta o cerrada. Para activar la configuración del tiempo de operación, presione el botón "Time" (6). En la pantalla "Time/Prog" (A-3) destella la indicación "m : s" o "h : m", dependiendo de la configuración previa. Para establecer el valor deseado, use la perilla de ajuste (1). Después de exceder 59 min. 50 seg., la indicación cambia automáticamente a "h : m". Después de exceder 99 horas 59 min., aparece la palabra "cont" en la pantalla "Time/Prog" (A-3). Esta operación continua solo puede interrumpirse al presionar el botón "Stop" (10). El conteo regresivo del tiempo inicia una vez que se alcanza la velocidad establecida. La pantalla siempre muestra el tiempo de operación restante. (Vea la figura 17.) Figura 17 3.3.3 Preselección de intensidad del freno y aceleración Esta función se activa con el botón "Accel/Decel" (5) (vea la Figura 18). Al presionar el botón una vez, destella la palabra "accel" (M7) en la pantalla "Acc/Dec" (A-2). La aceleración deseada puede preseleccionarse con la perilla de ajuste (1). El valor 0 equivale a la menor y el valor 9 a la mayor aceleración. Al presionar dos veces el botón "Accel/Decel" (5), la pantalla "Acc/Dec" (A-2) muestra la palabra "decel"(M8). Ahora puede preseleccionarse la intensidad del freno con la perilla de ajuste (1). El valor 9 equivale al más corto y el valor 0 al tiempo de freno más largo posible. Vea la "tabla 5: tiempos de aceleración y desaceleración" (APÉNDICE). En esta se muestran los tiempos de aceleración y desaceleración para las etapas de aceleración y desaceleración 0 a 9 para rotores permitidos. Figura 18 ES-10 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 3.3.4 Preselección de temperatura (solo para modelos refrigerados) Esta función se activa con el botón "Temp/Setup" (13). Después de presionar este botón, en la pantalla "Time/Prog" destella la indicación "°C" (A-4). Con la perilla (1) puede preseleccionarse la temperatura de prueba deseada en pasos de 1 °C en un rango de -20 °C hasta +40 °C. El valor se indica permanentemente en la pantalla (Figura 19) antes, durante y después de la operación. ¡Observe las mínimas temperaturas respectivas de los rotores a máxima velocidad Figura 19 3.3.5 Preenfriamiento (solo para modelos refrigerados) Si las muestras son sensibles a la temperatura, es útil preenfriar la centrífuga, el rotor y eventualmente los portatubos a la temperatura de trabajo requerida. Por lo tanto, inserte el rotor deseado y preajuste la temperatura respectiva. Al presionar en forma simultánea los botones "Temp/Setup" (13) (vea la Figura 20) y "Time" (6), inicia la operación. Durante la operación, la unidad selecciona automáticamente una velocidad rotacional que es equivalente al 20% de la velocidad rotacional permitida del rotor respectivo. Después de que se alcanza la temperatura predeterminada, puede salir de la operación de preenfriamiento con el botón "Stop" (10). Dependiendo del rotor insertado, el preenfriamiento dura aproximadamente entre 10 y 20 minutos. Figura 20 3.4 Corrección del radio Si usa adaptadores o reductores, puede cambiar el radio centrífugo del rotor respectivo. En ese caso, puede corregir el radio manualmente. Proceda de la siguiente manera: Cierre la tapa, luego presione el botón "Time" (6) (vea la Figura 21) y el botón "Prog/Setup" (11) al mismo tiempo y manténgalos presionados. En la pantalla "Time/Prog" (A-3) aparece la palabra "radius" (M9). Con la perilla de ajuste (1) puede preseleccionar la corrección del radio respectiva (vea la Tabla 7, APÉNDICE) en pasos de 0.1 cm. Una vez que haya establecido la corrección del radio, aparece la palabra "radius" (M9). Esta palabra será visible hasta que vuelva a poner la corrección del radio en 0. Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 3.5 Programa ES-11 Figura 21 3.5.1 Almacenamiento de programas Puede almacenar hasta 99 operaciones con todos los parámetros relevantes, incluyendo los rotores usados. Puede usar cualquier número de programa libre y volver a consultarlo. Coloque el rotor necesario en la centrífuga. Al presionar el botón "Prog/Setup" (11) en la pantalla "Time/Prog" (A-3) aparece la palabra "programm". Con la perilla de ajuste (1) puede seleccionar el número de programa deseado. Si un número de programa ya está ocupado, en la pantalla "RPM | RCF" (A-1) aparecerán las palabras "rotor" (M3) y "xx" (M4). En caso de números de programa libres, aparece 0. Figura 22 Cierre la tapa de la centrífuga. Ahora proceda como se describió anteriormente para configurar todos los parámetros de operación importantes. Si la tapa no se cierra al almacenar el programa, en la pantalla "RPM/RCF" (A-1) destellan las palabras "FirSt" y "CLOSE Lid" (vea la Figura 23) alternadamente. Si desea iniciar la operación sin almacenar el programa, en la pantalla "RPM/RCF" (A-1) destellan las palabras "First" y "PrESS StoreE" (vea la Figura 24) alternadamente. Figura 23 Figura 24 ES-12 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Para adopción de datos, presione el botón "Store" (12) (vea las Figuras 23 y 24) durante aproximadamente 1 segundo. Si el programa se almacena correctamente, aparece la palabra "StorE" en la pantalla "RPM/RCF" (A-1). Como resultado, desaparece la palabra "programm" (M10). Tan pronto como se suelta el botón "Store" (12), vuelve a aparecer la palabra "programm xx" (M10); la (xx) representa la ubicación de programa seleccionado. Si todos los números de programa están ocupados, puede tomar un número antiguo que ya no se necesite y simplemente ingresar los nuevos parámetros. 3.5.2 Recuperación de programas almacenados Para recuperar programas almacenados, presione el botón "Prog/Setup" (11) (vea la Figura 25) mientras la tapa está cerrada. En la pantalla "Time/Prog" (A-3) aparece "programm –"(M10). El número de programa deseado puede preseleccionarse con la perilla de ajuste (1). Los valores almacenados para ese programa aparecerán en las pantallas respectivas. Si dentro de la centrífuga está el rotor equivocado para el programa seleccionado, en la pantalla "RPM | RCF" (A1) destella la palabra "rotor" (M3). Al mismo tiempo, la palabra "FALSE" y el número de rotor almacenado " xx" (M4) destellan alternadamente. Figura 25 3.5.3 Salir del modo de programa Para salir del modo de programa, solo presione el botón "Prog/Setup" (11) (vea la Figura 25). Entonces, dentro de la pantalla "Time/Prog" aparece la palabra "programm". Configure la pantalla en "programm--" (M10) con la perilla de ajuste (1). 3.6 Arranque y paro de la centrífuga 3.6.1 Arranque de la centrífuga Puede arrancar la centrífuga ya sea con el botón "Start" (9) (vea la Figura 26) o con el botón "Quick Spin" (8). Con el botón "Start" (9), puede iniciar operaciones almacenadas u operaciones con parámetros preseleccionados manualmente. Una vez que finalice el tiempo de operación preseleccionado, la centrífuga parará automáticamente. Con el botón "Quick Spin" (8), puede iniciar operaciones que durarán solo unos segundos. Al presionar el botón "Quick Spin" (8), la centrífuga se acelera hasta las revoluciones preseleccionadas. En la pantalla "Time/Prog" (A-3), se indica el tiempo de operación transcurrido desde la fecha en que se presiona el botón "Quick Spin" (8). Al soltar el botón "Quick Spin" (8), la centrífuga se detiene y el tiempo de operación se indica hasta la apertura de la tapa. Figura 26 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-13 3.6.2 Botón "STOP" Con el botón "Stop" (10) (vea la Figura 27) puede interrumpir la operación en cualquier momento. Después de presionar el botón, la centrífuga se desacelera con la intensidad preseleccionada en descenso hasta parar completamente. Figura 27 3.7 Detección de desbalance En caso que el rotor no sea cargado de manera uniforme, el motor se apagará durante la aceleración. El rotor se desacelera hasta el paro total. Cuando en la pantalla "Time/Prog" (A-3) aparezca la palabra "error” (M11) junto con el número "01”, la diferencia de peso de las muestras es demasiado grande. Distribuya el peso de manera uniforme. Cargue el rotor como se describió en el capítulo 3.1.2 y 3.1.3. Cuando en la pantalla "Time/Prog" (A-3) aparezca la palabra "error” junto con el número "02" (vea la Figura 28), puede deberse a lo siguiente: el interruptor de desbalance está defectuoso. Figura 28 4. CONFIGURACIÓN 4.1 Ajustes básicos 4.1.1 Acceso al modo "Operating Data" Cuando use la centrífuga, se pueden configurar los siguientes parámetros: - Indicación de temperatura en °C o °F - Encendido/apagado de la señal acústica - Encendido/apagado del sonido del teclado - Preselección del volumen de la señal de sonido - Selección de melodía de la señal del sonido de final de operación o "end of run" Cuando la centrífuga está apagada, presione al mismo tiempo los botones "Time"(6) y "Door Open" (7) y encienda el interruptor principal de la centrífuga. Ahora suelte ambos botones y como resultado se ejecuta una prueba de pantalla durante aproximadamente 5 segundos. Todos los indicadores aparecerán al mismo tiempo (vea la Figura 29). ES-14 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figura 29 ATENCIÓN: Observe que debe ingresar en el programa como se describe en el punto 4.1.1 para cambiar los ajustes de los puntos 4.1.2 a 4.1.7. Una vez que haya guardado las configuraciones, puede cambiar al modo de programa normal nuevamente al apagar la centrífuga durante un tiempo breve. Todas las configuraciones cambiadas deben confirmarse con el botón "Start"(9). Aparece la palabra "Store"(12) en la pantalla "RPM | RCF"(A-1). ¡Solamente entonces son válidas las preselecciones! Figura 30 4.1.2 Indicación de temperatura Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón "Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) aparece la palabra "Service". Ahora seleccione la letra "C" con la perilla de ajuste (1). Como resultado, en la pantalla "RPM | RCF" (A-1) aparecen las palabras "CELSI/temp". Si presiona el botón "RPM | RCF" (4), la palabra "CELSI" destella y puede cambiar la pantalla a Fahrenheit "FAREN" con la perilla de ajuste (1) (vea la Figura 31). Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa normal al apagar la centrífuga durante un tiempo breve. Figura 31 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-15 4.1.3 Encendido y apagado de la señal Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón "Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) destella la palabra "Service". Ahora seleccione la letra "L" con la perilla de ajuste (1). Como resultado, aparecen las palabras "On Sound" en la pantalla "RPM | RCF" (4). Si presiona el botón "RPM | RCF" (4) ahora, destella la palabra "On" y puede apagar el sonido con la perilla de ajuste (1) (vea la Figura 32). Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa normal al apagar la centrífuga durante un tiempo breve. Figura 32 4.1.4 Preselección del volumen de la señal de sonido Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón "Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) destella la palabra "Service". Ahora seleccione la letra "U" con la perilla de ajuste (1). Como resultado, en la pantalla "RPM | RCF" (A-1) aparecen las palabras "Vol=09/Sound". Después de presionar el botón "RPM | RCF" (4), puede ajustar el volumen deseado entre 0 (bajo) y 9 (alto) con la perilla de ajuste (1) (vea la Figura 33). Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa normal al apagar la centrífuga durante un tiempo breve. Figura 33 4.1.5 Selección de melodía para la señal del sonido, final de operación Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón "Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) destella la palabra "Service". Ahora seleccione la letra "G" con la perilla de ajuste (1). Como resultado, en la pantalla "RPM | RCF" (A-1) aparece la palabra "SonGo/Sound". Después de presionar el botón "RPM | RCF" (4), puede seleccionar una melodía con la perilla de ajuste (1). (Vea la Figura 34.) Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa normal al apagar la centrífuga durante un tiempo breve. ES-16 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figura 34 4.1.6 Encendido y apagado del sonido del teclado Proceda como se describe en el punto 4.1.1 para ingresar en este modo de programa y luego presione el botón "Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) destella la palabra "Service". Ahora seleccione la letra "B" con la perilla de ajuste (1). Como resultado, en la pantalla "RPM | RCF" (A-1) aparece la palabra "ON/BEEP ". Después de presionar el botón "RPM | RCF" (4), puede encender el sonido del teclado (On) o apagarlo (Off) con la perilla de ajuste (1). (Vea la Figura 38.) Una vez que haya guardado las configuraciones (vea 4.1.1), puede cambiar nuevamente al modo de programa normal al apagar la centrífuga durante un tiempo breve. Figura 35 4.1.7 Extracción de datos operativos ATENCIÓN: Esto solo debe realizarlo un usuario avanzado o ingeniero de servicio. En el modo "Basic Adjustments" puede extraer los datos operativos de la centrífuga. Proceda como se describe en el punto 4.1.2 para ingresar en este modo del programa. Presione el botón "Accel/Decel" (5). En la pantalla "Acc/Dec" (A-2) destella la palabra "Service". El acceso a la diferente información se logra con la perilla de ajuste (1): A= arranques previos de la centrífuga H= horas de operación previa S= versión del software r= software del convertidor E= lista de mensajes de error previos h= tiempo de operación del motor La lista de los últimos 99 mensajes de error puede consultarse al presionar el botón "RPM | RCF" (4) y desplazarse en ella con la perilla de ajuste (1). Los códigos de error respectivos aparecen en la pantalla "RPM | RCF" (A-1). Consulte la "Tabla 6: mensajes de error" (vea APÉNDICE). Para regresar al modo de programa normal, apague la centrífuga durante un tiempo breve. Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-17 Figura 36 5. MANTENIMIENTO 5.1 Mantenimiento y limpieza 5.1.1 Cuidado general: El mantenimiento de la centrífuga se limita a mantener el rotor, la cámara del rotor y los accesorios del rotor limpios, así como lubricar regularmente los pernos de inserción del rotor de un rotor basculante (si está disponible). El lubricante más indicado es el aceite High TEF. No se permite el uso de lubricantes que contienen molycote y grafito. Preste atención especial a las partes de aluminio anodizado. La ruptura de los rotores puede ser causada incluso por el daño más ligero. En el caso de los rotores, portatubos o racks de tubos que entren en contacto con sustancias corrosivas, los puntos en cuestión deben limpiarse con cuidado. Algunos ejemplos de sustancias corrosivas son: álcalis, soluciones de jabón alcalinas, aminas alcalinas, ácidos concentrados, soluciones que contienen metales pesados, solventes clorados sin agua, soluciones salinas como agua con sal, fenol e hidrocarburos halogenados. Limpieza de unidades, rotores y accesorios: - Apague el dispositivo y desconéctelo de la corriente eléctrica antes de comenzar la limpieza o desinfección. No vierta líquidos en el interior de la caja. - No rocíe desinfectante sobre el dispositivo. - La limpieza profusa no solo tiene el objetivo sanitario sino también el de evitar la corrosión debido a la contaminación. - Para evitar el daño a las partes anodizadas como rotores, placas de reducción, etc., solo se deben usar detergentes de pH neutro con un pH de 6 a 8. No deben usarse agentes de limpieza alcalinos (pH > 8). - Después de la limpieza, asegúrese de secar todas las partes completamente, ya sea a mano o en un gabinete de aire caliente (temperatura máxima de +50 °C). - Es necesario recubrir las partes de aluminio anodizado con aceite contra la corrosión regularmente a fin de incrementar la vida útil y reducir la predisposición a la corrosión. - Debido a la humedad o a las muestras no selladas herméticamente, se puede formar condensado. El condensado debe eliminarse de la cámara del rotor con una tela suave regularmente. El procedimiento de mantenimiento debe repetirse cada 10 a 15 operaciones, o al menos una vez a la semana. - Conecte la unidad a la corriente eléctrica hasta después de que el equipo esté completamente seco. - No realice desinfección con rayos UV, beta o gamma ni otro tipo de radiación de energía. - Los rotores de metal pueden usarse en autoclave. - La tapa del rotor y adaptadores también pueden usarse en autoclave (máximo 121 °C, 20 minutos). - Los racks para tubos están hechas de PP y no pueden usarse en autoclave a 134 °C. ES-18 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 5.1.2 Limpieza y desinfección de la unidad 1. Abra la tapa antes de apagar la unidad. Desconéctela de la alimentación eléctrica. 2. Abra la tuerca del rotor al girar la llave del rotor en sentido contrario al reloj. 3. Retire el rotor. 4. Para la limpieza y desinfección de la unidad y la cámara del rotor, use el limpiador mencionado anteriormente. 5. Limpie todas las áreas accesibles del dispositivo y sus accesorios, incluyendo el cordón eléctrico, con una tela humedecida. 6. Lave los sellos de caucho y la cámara del rotor completamente con agua. 7. Frote los sellos de caucho con glicerol o talco para prevenir que se hagan quebradizos. Otros componentes de la unidad, como el seguro de la tapa, el eje del motor y el rotor no deben engrasarse. 8. Seque el eje del motor con una tela suave, seca y sin pelusa. 9. Controle la unidad y accesorios por posible daño. Asegúrese de que la centrífuga esté apagada y desconéctela de la corriente eléctrica. Luego remueva el polvo adherido de las ranuras de ventilación en la centrífuga con un cepillo suave. Haga esto al menos cada seis meses. 5.1.3 Limpieza y desinfección del rotor 1. Limpie y desinfecte los rotores, tapas de los rotores y adaptadores con el limpiador mencionado anteriormente. 2. Use un cepillo para botellas para limpiar y desinfectar los orificios internos del rotor. 3. Enjuague los rotores, tapa del rotor y adaptador con agua limpia. En particular, las perforaciones de los rotores angulares. 4. Para secar los rotores y accesorios, colóquelos en una toalla. Coloque los rotores angulares con los orificios hacia abajo. 5. Seque el cono del rotor con una tela suave, seca y sin pelusa e inspeccione por posible daño. No engrase el cono del rotor. 5.1.4 Desinfección de rotores de aluminio En caso de que material infeccioso se derrame dentro de la centrífuga, el rotor y cámara del rotor deben desinfectarse inmediatamente después de la operación. Los rotores pueden esterilizarse en autoclave a una temperatura máxima de 121 °C. 5.1.5 Desinfección de rotores de PP Uso en autoclave Tiempo recomendado en autoclave: 15 a 20 minutos a 121 °C (1 bar) ATENCIÓN: No debe excederse el tiempo de esterilización de 20 minutos. La esterilización repetida causará reducción de la resistencia mecánica del material plástico. Antes de introducirse en autoclave, el rotor de PP y adaptador deben limpiarse completamente para evitar que los residuos de suciedad se quemen dentro. Puede hacer caso omiso de las consecuencias de algunos residuos químicos en materiales plásticos a temperaturas ambiente. Sin embargo, a altas temperaturas durante la esterilización en autoclave, los residuos pueden corroer y destruir el plástico. Los objetos deben enjuagarse completamente con agua destilada después de limpiarse pero antes de la esterilización. Los residuos de cualquier líquido de limpieza pueden causar fisuras, blanqueamiento y manchas. Esterilización con gas Los adaptadores, botellas y rotores pueden esterilizarse con óxido de etileno. Asegúrese de ventilar los artículos después de la esterilización y antes de volver a usarlos. ATENCIÓN: Puesto que la temperatura puede aumentar durante la esterilización, los rotores, adaptadores y botellas no deben cerrarse y deben destornillarse completamente Esterilización química Las botellas, adaptadores y rotores pueden tratarse con los desinfectantes líquidos usuales. ATENCIÓN: Antes de aplicar cualquier otro método de limpieza o descontaminación que no sea el recomendado por el fabricante, póngase en contacto con este para asegurarse de que no dañará la unidad o el rotor. Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-19 5.1.6 Ruptura de vidrio Con altos valores g, el porcentaje de ruptura de tubos de vidrio se incrementa. Las astillas de vidrio deben retirarse inmediatamente del rotor, portatubos, adaptadores y de la cámara del rotor misma. Las astillas de vidrio finas rayarán y por lo tanto dañarán la cubierta superficial protectora de un rotor. Si permanecen astillas en la cámara del rotor, se acumulará polvo fino de metal debido a la circulación del aire. Este polvo metálico negro y muy fino contaminará significativamente la cámara del rotor, el rotor, los portatubos y las muestras. Si es necesario, remplace los adaptadores, tubos y accesorios para evitar más daños. Inspeccione regularmente los orificios internos del rotor por posibles residuos y daño. ATENCIÓN: Verifique las especificaciones relevantes de las centrífugas de tubos con el fabricante. 5.2 Vida útil de rotores, portatubos y accesorios Los rotores y tapa del rotor hechos de aluminio o acero inoxidable tienen una vida útil máxima de siete años a partir del primer uso. Las tapas y tapones transparentes de rotor hechos de PC o PP, así como los rotores, racks de tubos y adaptadores de PP tienen un tiempo operativo máximo de tres años a partir del primer uso. Condición para el tiempo operativo: condición de uso adecuado libre de daños, con el cuidado recomendado. 6. IDENTIFICACIÓN Y SOLUCIÓN DE PROBLEMAS 6.1 Mensaje de error: Causa / Solución Los mensajes de error se presentan para ayudar a localizar posibles errores más rápido. El diagnóstico al que se hace referencia en este capítulo puede no siempre ser el caso, ya que son solo errores y soluciones que ocurren teóricamente. Le pedimos que nos mantenga informados acerca de cualquier tipo de error que ocurra y que no esté descrito en este capítulo. Solo mediante su información podremos mejorar este manual de operación. Muchas gracias de antemano por su apoyo. 6.2 Estudio de posibles mensajes de error y sus soluciones 6.2.1 Apertura de la tapa durante una falla de la corriente eléctrica (Apertura de emergencia de la tapa) En caso de falla de la corriente eléctrica o mal funcionamiento, la tapa de la centrífuga puede abrirse manualmente para recuperar las muestras. En los modelos FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (cerradura accionada por motor) Proceda de la siguiente manera: 1. Apague la centrífuga y desenchufe el cordón eléctrico y espere hasta que el rotor se haya detenido completamente (esto puede durar varios minutos). 2. En el lado izquierdo de la caja de la centrífuga hay un detenedor de plástico. Retire este detenedor y detrás de él hay una tuerca hexagonal. 3. Tome la llave incluida, colóquela en el agujero y bloquee la llave con la tuerca hexagonal (vea la Figura 37). 4. Ahora gire la llave a la derecha (en sentido del reloj) hasta el límite. ATENCIÓN: a) Solo gire hasta el límite; no apriete la tuerca. b) Ahora abra la tapa de la centrífuga. c) Encienda la centrífuga nuevamente para reanudar el trabajo. ES-20 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figura 37 For 5714 Realice lo siguiente (ver la Figura 38): ATENCIÓN: • Apague la centrífuga y desconecte el cable de alimentación, espere hasta que el rotor se detenga completamente (esto puede demorar varios minutos) En el lado derecho de la centrífuga hay un detenedor plástico (Figura 38). Quite este detenedor, que está conectado a la cerradura de la tapa, horizontalmente desde la carcasa hasta que la tapa de la centrífuga se abra. •Ahora abra la tapa de la centrífuga Figura 38 6.2.2 Descripción del sistema de mensajes de error El mensaje de error "error" (M11) se muestra en la pantalla "Time/Prog" (A-3) (vea la Figura 39). La información detallada acerca de los posibles mensajes de error está en la "tabla 6: mensajes de error" (vea Apéndice). Figura 39 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-21 7. RECEPCIÓN DE CENTRÍFUGAS PARA REPARACIÓN Riesgo para la salud por equipo, rotores y accesorios contaminados. En caso de enviar la centrífuga para reparación al fabricante, observe lo siguiente: La centrífuga debe descontaminarse y limpiarse antes del envío para protección de las personas, medio ambiente y material. Certificado de descontaminación en la entrega de devolución de bienes (vea APÉNDICE) Nos reservamos el derecho de no aceptar centrífugas contaminadas. Todos los costos adicionales incurridos para la limpieza y desinfección de las unidades s e agregarán al débito de la cuenta del cliente 8. TRANSPORTE, ALMACENAMIENTO Y ELIMINACIÓN 8.1 Transporte Antes del transporte, retire el rotor. Solo transporte la unidad en el embalaje original. Use un auxiliar de transporte para el transporte a través de largas distancias para fijar el eje del motor. Transporte general Temperatura del aire -25 a 60 ° Humedad relativa 10 a 75 % Presión de aire 30 a 106 kPa 8.2 Almacenamiento Durante el almacenamiento de la centrífuga, se deben observar las siguientes condiciones ambientales: En empaquetado de transporte Temperatura del aire -25 a 55 ° Humedad relativa 10 a 75 % Presión de aire 70 a 106 kPa 8.3 Transporte, instalación, traslado y desecho de la centrífuga FC5714/FC5718/FC5718R/FC5816/FC5816R Estas instrucciones complementan las anteriores instrucciones generales del capítulo 8 pero no las sustituyen. 8.3.1 Transporte • Transporte el dispositivo en su embalaje original • Dos personas deben trasportar siempre la centrífuga. • Para largas distancias, utilice la ayuda de un transporte. 8.3.2 Instalación Dos personas deben trasportar siempre la centrífuga.  Abrir la caja y sacar el dispositivo. 1. Corte la cinta adhesiva. 2. Abra las 4 solapas de la caja. 3. Saque los accesorios. 4. Introduzca las manos hasta colocarlas debajo del dispositivo y, con ayuda de otra persona, saque la centrífuga de la caja. ES-22 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ADVERTENCIA: Peligro de levantamiento. Existe riesgo de lesiones si una sola persona la levanta. Solicite ayuda al levantar o mover el equipo.  Colocar el dispositivo en una mesa de laboratorio estable, horizontal y no resonante. 1. Retire el material protector para el transporte delantero y trasero. 2. Retire la funda de plástico. 3. Mantenga una distancia mínima de 30 cm entre los laterales del equipo y los dispositivos adyacentes y entre la parte trasera y la pared. 4. Instale el dispositivo en una ubicación con buena ventilación que esté protegida del sol directo para evitar que se recaliente.  Conectar el dispositivo 1. Después de realizar la instalación, espere durante cuatro horas para encender la centrífuga para evitar dañar el compresor. 2. Compruebe que la tensión y la frecuencia de la red eléctrica se ajustan a los requisitos de la placa de características del dispositivo (consulte la parte trasera del dispositivo) y después conecte el dispositivo a la red eléctrica.  Retirar el material protector para el transporte de la cámara del rotor. 1. Encienda el dispositivo con el interruptor de encendido. 2. Abra la tapa de la centrífuga utilizando el botón de apertura. 3. Retire el material protector para el transporte. 4. Coloque el rotor de forma vertical en el eje del motor. 5. Gire la tuerca del rotor con la llave incluida en el sentido de las agujas del reloj hasta que quede apretada.  Ya puede utilizar el dispositivo. Conserve el embalaje y todo el material protector para el transporte para mover el dispositivo en un futuro. 8.3.3 Embalaje Embale la centrífuga siguiendo los pasos anteriores en el sentido contrario. 8.3.4 Entrega del dispositivo Al pasar el equipo a terceros, por favor asegúrese de incluir también este manual de instrucciones. Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9. FICHA TÉCNICA 9.1 Especificaciones 9.1.1 Centrifuge FC5714 Modelo Rango de v elocidad Máxima RCF Máxima capacidad (rotor) Rango de temperatura Tiempo de operación Niv el de ruido (dependiendo del rotor) Densidad permitida a máxima v elocidad Energía cinética permitida Conexión CA de corriente eléctrica principal FC5714 200 rpm -14000 rpm;10 rpm/set 18624 x g;10 x g/set 4x100ml Enfriamiento por aire 10 seg a 99 hr 59 min 59 seg o continuo ≤ 63 ± 2 dB(A) 1.2 g/ml 5595 Nm 230 V ~ 50/60 Hz ± 10 % 1.3 A 240 W Fluctuación de v oltaj e Consumo de corriente Consumo de energía Dimensiones (Ancho × Profundidad × Alto) Peso neto (sin rotor) 362 x 493 x 330 mm 14.3 x 19.4 x 13.0 in Dimensiones de env ío (Ancho × Profundidad x Alto) 30 kg Peso de env ío (sin rotor) 66 lb Condiciones ambientales (EN/IEC 61010-1) Ambiente Altitud Temperatura ambiente Máx. humedad relativ a Categoría de sobrev oltaj e (IEC 60364-4-443) Grado de contaminación Clase de protección 230 V ~ 50/60 Hz ± 10 % 1.3 A 240 W para uso interno solamente Use hasta a una altitud de 2000 m 2 °C hasta 35 °C Máx. humedad relativa 80 % para temperaturas de hasta 31 °C, disminuyendo linealmente a 50% de humedad relativa hasta 35 °C. No adecuada para usarse en ambientes peligrosos. II 2 I ES-23 ES-24 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.2 Centrífuga FC5718 Modelo Rango de v elocidad Máxima RCF Máxima capacidad (rotor) Rango de temperatura Tiempo de operación Niv el de ruido (dependiendo del rotor) Densidad permitida a máxima v elocidad Energía cinética permitida Conexión CA de corriente eléctrica principal Fluctuación de v oltaj e Consumo de corriente Consumo de energía Dimensiones (Ancho × Profundidad × Alto) FC5718 200 rpm -18000 rpm;10 rpm/set 23542 x g;10 x g/set 4x100ml Enfriamiento por aire 10 seg a 99 hr 59 min 59 seg o continuo ≤ 60 ± 2 dB(A) 1.2 g/ml 16672 Nm 230 V ~ 50/60 Hz 230 V ~ 50/60 Hz ± 10 % ± 10 % 2.0 A 2.0 A 455 W 455 W 408 x 499 x 351 mm 16.1 x 19.7 x 13.8 in Peso neto (sin rotor) Dimensiones de env ío (Ancho × Profundidad x Alto) 43 kg 95 lb Peso de env ío (sin rotor) Condiciones ambientales (EN/IEC 61010-1) Ambiente Altitud Temperatura ambiente Máx. humedad relativ a para uso interno solamente Use hasta a una altitud de 2000 m 2 °C hasta 35 °C Máx. humedad relativa 80 % para temperaturas de hasta 31 °C, disminuyendo linealmente a 50% de humedad relativa hasta 35 °C. Categoría de sobrev oltaj e (IEC 60364-4-443) Grado de contaminación Clase de protección No adecuada para usarse en ambientes peligrosos. II 2 I Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.3 Centrifuge FC5718R Modelo Rango de v elocidad Máxima RCF Máxima capacidad (rotor) Rango de temperatura Tiempo de operación Niv el de ruido (dependiendo del rotor) Densidad permitida a máxima v elocidad Energía cinética permitida Conexión CA de corriente eléctrica principal Fluctuación de v oltaj e FC5718R 200 rpm -18000 rpm;10 rpm/set 23542 x g;10 x g/set 4x100ml -20~40℃ 10 seg a 99 hr 59 min 59 seg o continuo ≤ 60 ± 2 dB(A) 1.2 g/ml 25111 Nm 230 V ~ 50/60 Hz 230 V ~ 50/60 Hz ± 10 % 3.0 A 660 W Consumo de corriente Consumo de energía Dimensiones (Ancho × Profundidad × Alto) ± 10 % 3.0 A 660 W 407 x 731 x 359 mm 16.0 x 28.8 x 14.1 in Peso neto (sin rotor) Dimensiones de env ío (Ancho × Profundidad x Alto) 60 kg 132 lb Peso de env ío (sin rotor) Condiciones ambientales (EN/IEC 61010-1) Ambiente Altitud Temperatura ambiente Máx. humedad relativ a para uso interno solamente Use hasta a una altitud de 2000 m 2 °C hasta 35 °C Máx. humedad relativa 80 % para temperaturas de hasta 31 °C, disminuyendo linealmente a 50% de humedad relativa hasta 35 °C. Categoría de sobrev oltaj e (IEC 60364-4-443) Grado de contaminación Clase de protección No adecuada para usarse en ambientes peligrosos. II 2 I ES-25 ES-26 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.4 Centrifuge FC5816 Modelo Rango de v elocidad Máxima RCF Máxima capacidad (rotor) Rango de temperatura Tiempo de operación Niv el de ruido (dependiendo del rotor) Densidad permitida a máxima v elocidad Energía cinética permitida Conexión CA de corriente eléctrica principal Fluctuación de v oltaj e Consumo de corriente FC5816 200 rpm -15000 rpm;10 rpm/set 21379 x g;10 x g/set 6 x 250ml Enfriamiento por aire 10 seg a 99 hr 59 min 59 seg o continuo ≤ 61 ± 2 dB(A) 1.2 g/ml 34363 Nm 230 V ~ 50/60 Hz 230 V ~ 50/60 Hz ± 10 % 2.4 A 530 W Consumo de energía ± 10 % 2.4 A 530 W Dimensiones (Ancho × Profundidad × Alto) 446 x 538 x 354 mm Peso neto (sin rotor) 17.6 x 21.2 x 13.9 in Dimensiones de env ío (Ancho × Profundidad x Alto) 52 kg 115 lb Peso de env ío (sin rotor) Condiciones ambientales (EN/IEC 61010-1) Ambiente Altitud Temperatura ambiente Máx. humedad relativ a para uso interno solamente Use hasta a una altitud de 2000 m 2 °C hasta 35 °C Máx. humedad relativa 80 % para temperaturas de hasta 31 °C, disminuyendo linealmente a 50% de humedad relativa hasta 35 °C. Categoría de sobrev oltaj e (IEC 60364-4-443) Grado de contaminación Clase de protección No adecuada para usarse en ambientes peligrosos. II 2 I Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.5 Centrifuge FC5816R Modelo Rango de v elocidad Máxima RCF Máxima capacidad (rotor) Rango de temperatura Tiempo de operación Niv el de ruido (dependiendo del rotor) Densidad permitida a máxima v elocidad Energía cinética permitida Conexión CA de corriente eléctrica principal Fluctuación de v oltaj e Consumo de corriente Consumo de energía Dimensiones (Ancho × Profundidad × Alto) FC5816R 200 rpm -16000 rpm;10 rpm/set 24325 x g;10 x g/set 6 x 250ml -20~40℃ 10 seg a 99 hr 59 min 59 seg o continuo ≤ 63 ± 2 dB(A) 1.2 g/ml 34363 Nm 230 V ~ 50/60 Hz 230 V ~ 50/60 Hz ± 10 % ± 10 % 3.7 A 3.7 A 785 W 785 W 723 x 538 x 354 mm 28.5 x 21.2 x 13.9 in Peso neto (sin rotor) Dimensiones de env ío (Ancho × Profundidad x Alto) 77 kg Peso de env ío (sin rotor) 170 lb Condiciones ambientales (EN/IEC 61010-1) Ambiente Altitud Temperatura ambiente Máx. humedad relativ a para uso interno solamente Use hasta a una altitud de 2000 m 2 °C hasta 35 °C Máx. humedad relativa 80 % para temperaturas de hasta 31 °C, disminuyendo linealmente a 50% de humedad relativa hasta 35 °C. Categoría de sobrev oltaj e (IEC 60364-4-443) Grado de contaminación Clase de protección No adecuada para usarse en ambientes peligrosos. II 2 I ES-27 ES-28 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.6 Centrifuge FC5916 Model FC5916 200 rpm -16000 rpm;10 rpm/set Speed Range 24325 x g;10 x g/set Maximum RCF Maximum Capacity(Rotor) 4x750ml Temperature range(N/A) Air cool Running Time 10 sec to 99 hr 59 min 59 sec or continuous Noise lev el (depending on the rotor) Allow able density at maximum speed Allow able kinetic energy Mains pow er connection AC Voltage fluctuation ≤ 63 ± 2 dB(A) 1.2 g/ml 60629 Nm 230 V ~ 50/60 Hz ± 10 % Current consumption 2.8 A Pow er consumption 640 W 120 V ~ 50/60 Hz ± 10 % 5.6 A 680 W 540 x 670 x 390 mm Dimensions (W × D × H) 21.3 x 26.4 x 15.4 in 85 kg Net Weight (w ithout rotor) 187 lb 780 x 670 x 590 mm Shipping Dimensions (W × D × H) 30.7 x 26.4 x 23.3 in 98 kg Shipping Weight (w ithout rotor) 216 lb Ambient conditions (EN/IEC 61010-1) For indoor use only Env ironment Use up to an altitude of 2000 m Altitude 2°C up to 35 °C Ambient temperature Max. relative humidity 80 % for temperatures up to 31°C, Max. relative humidity decreasing linearly to 50 % relative humidity up to 35°C. Ov ervoltage category (IEC 60364-4443) Degree of contamination Class of protection Not suitable for use in hazardous env ironments. II 2 I Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.7 Centrifuge FC5916R Model FC5916R 200 rpm -16000 rpm;10 rpm/set Speed Range 26331 x g;10 x g/set Maximum RCF Maximum Capacity(Rotor) 4x750ml Temperature range(Digital) -20~40℃ Running Time 10 sec to 99 hr 59 min 59 sec or continuous Noise lev el (depending on the rotor) Allow able density at maximum speed Allow able kinetic energy ≤ 63 ± 2 dB(A) 1.2 g/ml 54458 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % Current consumption 7.2 A 20 A Pow er consumption 1630 W 1750 W Mains pow er connection AC Voltage fluctuation 730 x 670 x 390 mm Dimensions (W × D × H) 28.7 x 26.4 x 15.4 in 118 kg Net Weight (w ithout rotor) 260 lb 900 x 750 x 560 mm Shipping Dimensions (W × D × H) 40.0 x 29.5 x 22.0 in 137 kg Shipping Weight (w ithout rotor) 137 kg Ambient conditions (EN/IEC 61010-1) for indoor use only Env ironment Use up to an altitude of 2000 m Altitude 2°C up to 35 °C Ambient temperature Max. relative humidity 80 % for temperatures up to 31°C, Max. relative humidity decreasing linearly to 50 % relative humidity up to 35°C. Ov ervoltage category (IEC 60364-4443) II Degree of contamination 2 Class of protection I Not suitable for use in hazardous env ironments. ES-29 ES-30 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.2 Diagramas y dimensiones D H W D H W Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Modelo FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R W (mm / in.) 362 / 14.3 408 / 16.1 408 / 16.1 446 / 17.6 723 / 28.5 540 / 21.3 730 / 28.7 D (mm / in.) 493 / 19.4 499 / 19.7 731 / 28.8 538 / 21.2 538 / 21.2 670 / 26.4 670 / 26.4 H (mm / in.) 330 / 13.0 351 / 13.8 359 / 14.1 354 / 13.9 354 / 13.9 390 / 15.4 390 / 15.4 10. CONFORMIDAD La marca correspondiente en el producto indica la conformidad con los siguientes estándares. Marca Estándar Este producto es conforme con las normas armonizadas aplicables de directivas de la UE 2011/65/UE (RoHS) 2014/30/UE (EMC) y 2014/35/UE (LVD). La declaración de conformidad de la UE está disponible en línea en www.ohaus.com/ce. Desecho En conformidad con la Directiva Europea 2012/19/UE sobre Residuos de Equipos Eléctricos y Electrónicos (WEEE), este dispositivo no puede desecharse con la basura doméstica. Esto también es aplicable para países fuera de la UE, según sus requerimientos específicos. Deseche este producto de acuerdo con las regulaciones locales en el punto de recolección especificado para equipos eléctricos y electrónicos. Si tiene alguna pregunta, comuníquese con la autoridad responsable o con el distribuidor a quien compró este dispositivo. En caso de que este dispositivo sea transferido a otras partes (para uso privado o profesional), también deberá mencionarse el contenido de esta regulación. Para obtener instrucciones de eliminación en Europa, consulte www.ohaus.com/weee. Gracias por su contribución a la protección ambiental. ES-31 ES-32 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11. APÉNDICE TABLA 1: PESO NETO PERMITIDO TABLA 2: TEMPERATURAS MÍNIMAS A MÁ XIMA VELOCIDAD TABLA 3: MÁ XIMA VELOCIDAD Y VALORES RCF PARA ROTORES PERMITIDOS TABLA 4: TIEMPOS DE ACELERACIÓ N Y DESACELERACIÓ N TABLA 5: MENSAJES DE ERROR TABLA 6 CORRECCIÓ N DEL RADIO TABLA 7: FORMULARIO DE DEVOLUCIÓ N / CERTIFICADO DE DESCONTAMINACIÓ N 11.1 Table 1: PESO NETO PERMITIDO FC5916R FC5916 FC5816R FC5816 Peso neto permitido FC5718R Descripció n del rotor FC5718 Nú mero del rotor FC5714 Velocidad má x. ( x1000 rpm) 14.0 14.0 15.0 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 24x 3.4 g 15.0 16.0 16.0 16.0 30314820 Rotor, Sw ing out, 4x 250ml, ID 4.5 4x 536 g 4.5 8.0 30314821 Rotor, Angle, 6x 250ml FB, ID 8.0 6x 355 g 8.0 8.0 4.5 30314822 Rotor, Sw ing out, 4x 100ml, ID 4x 340 g 4.5 4.5 4.0 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 4x 465 g 5.0 5.0 4.5 4.5 4.5 30314824 Rotor, Sw ing out, 2x 3MTP, ID 2x 310 g 4.5 4.5 4.5 4.5 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 6x 140 g 11.0 13.5 9.0 9.0 30314826 Rotor, Angle, 6x 85ml RB, ID 11.0 6x 140 g 13.0 11.0 11.0 12.0 12.0 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 4x 140 g 12.0 12.0 15.0 15.0 30314828 Rotor, Sw ing out, 16x 50ml, ID 4x 557 g 4.5 4.5 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-33 7.5 7.5 30314829 Rotor, Angle, 10x 50ml FA, ID 9.0 10x 74 g 10.5 10.0 10.0 6.0 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 6x 72 g 6.0 6.0 12.0 12.0 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 13.0 6x 94 g 13.0 13.0 13.0 4.5 4.5 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30x 32 g 4.5 4.5 4.5 12.0 12.0 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 12.0 20x 18 g 12.0 12.0 12.0 6.0 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 12x 25 g 6.0 6.0 13.5 13.5 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 15.0 44x 3.8 g 16.0 16.0 16.0 12.0 13.0 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30x 3.4 g 14.0 15.0 15.0 14.0 14.0 15.0 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 24x 3.4 g 15.0 16.0 15.0 15.0 30314839 30314840 Rotor, Angle, 12x 1.5/2.0ml, ID Rotor, Angle, 64x 0.5ml, ID 12x 3.4 g 64x 2.3 g 18.0 18.0 13.5 13.5 15.0 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 4x 3.5 g 15.0 15.0 15.0 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID 4x 995 g 4x 708 g 4.0 4.0 8.0 8.0 ES-34 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.2 Table 2: TEMPERATURAS MÍNIMAS A MÁXIMA VELOCIDAD Nú mero de rotor Descripció n del rotor 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 30314840 Rotor, Angle, 64x 0.5ml, ID 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID Modelo FC5718R FC5816R FC5916R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5916R FC5916R Velocidad má xima 15000 rpm 15000 rpm 16000 rpm 8000 rpm 4500 rpm 4500 rpm 5000 rpm 5000 rpm 4500 rpm 4500 rpm 4500 rpm 13500 rpm 13500 rpm 9000 rpm 9000 rpm 13000 rpm 12000 rpm 12000 rpm 16000 rpm 7500 rpm 7500 rpm 10500 rpm 6000 rpm 6000 rpm 12000 rpm 12000 rpm 13000 rpm 4500 rpm 4500 rpm 12000 rpm 12000 rpm 12000 rpm 6000 rpm 6000 rpm 13500 rpm 13500 rpm 16000 rpm 14000 rpm 14000 rpm 15000 rpm 15000 rpm 15000 rpm 16000 rpm 18000 rpm 18000 rpm 13500 rpm 13500 rpm 15000 rpm 15000 rpm 15000 rpm 4500 rpm 8000 rpm N-max 6 °C 6 °C 1 °C -1 ° C -7 ° C -7 ° C -3 ° C -3 ° C -6 ° C -6 ° C -8 ° C 17 ° C 17 ° C -5 ° C -5 ° C 8 °C 5 °C 5 °C 7 °C 9 °C 9 °C 2 °C -8 ° C -8 ° C -5 ° C -5 ° C -4 ° C -9 ° C -9 ° C 2 °C 2 °C -2 ° C -11.5 ° C -11.5 ° C 1 °C 1 °C 3 °C 9 °C 9 °C 3 °C 6 °C 6 °C 1 °C -2 ° C -2 ° C 2 °C 2 °C -1 ° C -1 ° C -2 ° C 4 °C 0 °C Todas las indicaciones de temperatura se refieren a una temperatura ambiente de 23 ° C. Al exceder este valor o la radiación solar directa a la centrífuga, estos valores no pueden mantenerse. Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-35 11.3 Table 3: MÁXIMA VELOCIDAD Y VALORES RCF PARA ROTORES PERMITIDOS Velocidad má x. Nú mero del rotor 30304361 30314820 30314821 30314822 30314823 30314824 30314825 30314826 30314827 30314828 Descripció n del rotor Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable Rotor, Sw ing out, 4x 250ml, ID Rotor, Angle, 6x 250ml FB, ID Rotor, Sw ing out, 4x 100ml, ID Rotor, Sw ing out, 4x 100ml, ID, Sealable Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID Rotor, Angle, 4x 85ml RB, ID, Hi Rotor, Sw ing out, 16x 50ml, ID Modelo RPM RCF FC5714 14000 18624 x g FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5816 4500 3780 x g FC5816R 4500 3780 x g FC5816 8000 10016 x g FC5816R 8000 10016 x g FC5916 8000 10016 x g FC5916R 8000 10016 x g FC5714 4500 3350 x g FC5718 4500 3350 x g FC5718R 4500 3350 x g FC5714 4000 2486 x g FC5718 5000 3885 x g FC5718R 5000 3885 x g FC5714 4500 2716 x g FC5718 4500 2716 x g FC5718R 4500 2716 x g FC5816 4500 2716 x g FC5816R 4500 2716 x g FC5916 4500 2716 x g FC5916R 4500 2716 x g FC5718 11000 13932 x g FC5718R 13500 20984 x g FC5718 9000 10413 x g FC5718R 9000 10413 x g FC5816 11000 15555 x g FC5816R 13000 21726 x g FC5916 11000 15555 x g FC5916R 13000 21726 x g FC5718 12000 14809 x g FC5718R 12000 14809 x g FC5816 12000 14809 x g FC5816R 12000 14809 x g FC5916 15000 23140 x g FC5916R 16000 26328 x g FC5816 4500 3735 x g FC5816R 4500 3735 x g ES-36 30314829 30314830 30314831 30314832 30314833 30314834 30314835 30314836 30314838 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Rotor, Angle, 20x 10ml RB, ID, Hi Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable Rotor, Angle, 24x 1.5/2.0ml, ID FC5718 7500 8174 x g FC5718R 7500 8174 x g FC5816 9000 11771 x g FC5816R 10500 16022 x g FC5916 10000 14532 x g FC5916R 10500 16022 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 12000 13522 x g FC5718R 12000 13522 x g FC5816 13000 15869 x g FC5816R 13000 15869 x g FC5916 13000 15869 x g FC5916R 13000 15869 x g FC5714 4500 2830 x g FC5718 4500 2830 x g FC5718R 4500 2830 x g FC5816 4500 2830 x g FC5816R 4500 2830 x g FC5718 12000 15775 x g FC5718R 12000 15775 x g FC5816 12000 15775 x g FC5816R 12000 15775 x g FC5916 12000 15775 x g FC5916R 12000 15775 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 13500 17113 x g FC5718R 13500 17113 x g FC5816 15000 21128 x g FC5816R 16000 24039 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5714 12000 15131 x g FC5718 13000 17758 x g FC5718R 14000 20595 x g FC5916 15000 23643 x g FC5916R 15000 23643 x g FC5714 14000 18624 x g FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314839 30314840 30314841 30553085 30553086 Rotor, Angle, 12x 1.5/2.0ml, ID Rotor, Angle, 64x 0.5ml, ID Rotor, Angle, 4x 8-w PCR Strip, ID Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID ES-37 FC5718 18000 23542 x g FC5718R 18000 23542 x g FC5718 13500 16298 x g FC5718R 13500 16298 x g FC5718 15000 15343 x g FC5718R 15000 15343 x g FC5916 15000 15343 x g FC5916R 15000 15343 x g FC5916 4000 3434 x g FC5916R 4500 4346 x g FC5916 8000 10367 x g FC5916R 8000 10367 x g ES-38 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.4 Table 4: TIEMPOS DE ACELERACIÓN Y DESACELERACIÓN Nú mero de producto Descripció n del rotor Modelo Aceleració n* Tiempo en seg. Nivel 0 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314820 Rotor, Sw ing out, 4x 250ml, ID 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314828 Rotor, Sw ing out, 16x 50ml, ID 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R 200 240 230 251 249 204 204 309 309 664 664 573 573 110 100 100 110 100 100 230 160 160 452 452 249 249 400 500 360 360 697 2313 463 549 300 300 506 506 448 448 311 311 380 380 753 753 480 480 101 89 90 360 360 446 446 264 264 Nivel 9 25 23 25 25 20 21 21 34 34 130 130 66 66 14 15 15 14 15 15 25 20 20 43 43 27 27 40 60 40 40 85 70 48 69 60 60 60 60 50 50 34 34 63 62 115 115 60 60 11 13 12 40 40 48 48 28 28 Desaceleració n* Tiempo en seg. Nivel 0 230 240 420 610 635 421 421 458 458 2906 2906 1903 1903 170 150 150 170 150 150 340 360 360 616 616 488 488 960 1260 1050 1050 2313 1630 1654 1307 820 820 1745 1745 1251 1251 387 387 796 1069 2395 2395 1747 1747 206 239 463 570 570 1323 1323 921 921 Nivel 9 35 16 17 26 27 30 30 36 36 92 92 84 84 19 15 15 17 15 15 26 15 15 38 38 23 23 40 50 40 40 70 76 46 67 40 40 49 49 45 45 36 36 37 36 72 72 68 68 14 11 11 30 30 49 49 32 32 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314832 Rotor, Angle, 30x15ml RB/FA, ID 30314833 Rotor, Angle, 20x10ml RB, ID, Hi 30314834 Rotor, Angle, 12x15ml RB/FA, ID 30314835 Rotor, Angle, 44x1.5/2.0ml, ID 30314836 Rotor, Angle, 30x1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x1.5/2.0ml, ID 30314839 Rotor, Angle, 12x1.5/2.0ml, ID 30314840 Rotor, Angle, 64x0.5ml, ID 30314841 Rotor, Angle, 4x8-w PCR Strip, ID 30553085 Rotor SwingOut 4x750ml D99mm ID SL 30553086 Rotor Angle 4x500ml D69mm ID * Nota: acelera de 0 a Vmax; desacelera de Vmax a 0. FC5714 FC5718 FC5718R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5916 FC5916R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5718 FC5718R FC5916 FC5916R FC5916 FC5916R FC5916 FC5916R 157 160 160 149 149 360 360 512 512 305 305 102 88 88 196 196 257 274 233 233 250 200 210 221 221 200 210 230 251 249 204 204 240 240 150 150 130 130 100 100 483 551 575 575 ES-39 19 15 15 25 25 40 40 54 54 32 32 11 13 12 22 22 29 31 25 25 27 25 25 23 23 22 23 25 25 30 21 21 25 25 17 17 16 16 12 12 47 55 73 73 370 380 380 985 985 570 570 1439 1439 988 988 167 222 433 204 204 436 485 283 500 280 360 360 561 561 230 240 420 610 635 421 421 210 210 170 170 130 130 201 201 1287 1501 2317 2317 20 10 10 20 20 30 30 47 47 37 37 14 11 11 18 18 28 29 30 25 34 25 30 30 30 35 16 17 26 27 30 30 20 20 12 12 12 12 12 12 49 54 82 82 ES-40 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.5 Tabla 5: Mensajes de error Error-No.: 1 2 4 8 11 12 14 Tapa CERRADA 33 34 38 40 41 42 43 44 45 46 47 48 49 55 70 99 FALSO No. de rotor Descripció n Aparece desbalance Sensor de desbalance defectuoso Interruptor de desbalance activado por má s de 5 segundos Transpondedor en rotor defectuoso Sensor de temperatura defectuoso Exceso de temperatura de la cá mara Salto de velocidad demasiado grande entre dos mediciones Tapa abierta mientras el motor está funcionando Contacto de la tapa defectuoso Motor de la tapa bloqueado Comunicació n con convertidor de frecuencia alterada durante el arranque Comunicació n con convertidor de frecuencia alterada durante el paro Cortocircuito en el convertidor de frecuencia Bajo voltaje de convertidor de frecuencia Alto voltaje del convertidor de frecuencia Exceso de temperatura del convertidor de frecuencia Exceso de temperatura del motor Exceso de corriente del convertidor de frecuencia Tiempo excedido entre la unidad de control y el controlador de frecuencia Otro error del convertidor de frecuencia Exceso de velocidad Tiempo excedido entre el controlador y la interfaz RS232 Rotor no permitido en esta centrífuga El rotor insertado no existe en el programa Rotor no detectado Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-41 11.6 Table 6: CORRECCIÓN DEL RADIO Compatibilidad FC5916 FC5916R ● ● ● ● ● ● ● -0.3 ● ● ● ● ● ● ● -1.0 ● ● ● ● ● ● ● ● ● -1.0 ● ● 15.9 -0.8 ● ● 16.3 -0.4 ● ● Rack, 3x 50ml D29mm FA, 2/pk 16.7 0.0 ● ● 30314907 Rack, 1x 50ml D34mm RB, 2/pk 15.8 -0.9 ● ● 30314908 Rack, 2x 50ml D29mm Rim, 2/pk 15.8 -0.9 ● ● 30314909 Rack, 1x 50ml D29mm Rim, 2/pk 16.1 -0.6 ● ● 30314910 Rack, 5x 25ml D24.5mm RB, 2/pk 16.1 -0.6 ● ● 30314911 Rack, 3x 25ml D24.5mm Rim, 2/pk 15.5 -1.2 ● ● 30314912 Rack, 9x 15ml D17mm RB, 2/pk 16.3 -0.4 ● ● 30314913 Rack, 7x 15ml D17mm FA, 2/pk 16.4 -0.3 ● ● 30314914 Rack, 7x 15ml D17mm Rim, 2/pk 16.1 -0.6 ● ● 30314915 Rack, 8x 10ml D16mm RB, 2/pk 16.3 -0.4 ● ● 30314916 Rack, 14x 5-7ml D13mm RB, 2/pk 15.9 -0.8 ● ● 30314917 Rack, 10x 1.6-7ml D13mm Vac, 2/pk 15.9 -0.8 ● ● 30304367 Rack, 8x 4-10ml D16mm Vac, 2/pk 16.3 -0.4 ● ● 30304368 Adapter, 9x 1.5/2.0ml D11mm, 2/pk 15.7 -1.0 ● ● Blank NO Bucket/Adapter 14.1 ● ● ● ● 30304369 Adapter, 8x 1.5ml D11mm, 2/pk 13.2 -0.9 ● ● ● ● 13.3 -0.8 ● ● ● ● Blank NO Bucket/Adapter 8.5 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 Blank NO Bucket/Adapter 16.7 30314903 Rack, 1x 250ml D62mm FB, 2/pk 15.7 30314904 Rack, 1x 100ml D41mm RB, 2/pk 30314905 Rack, 3x 50ml D29mm RB, 2/pk 30314906 30304370 30314821 Rotor, Angle, 6x 250ml FB, ID 30304371 Adapter, 4x 15ml D17mm FA, 2/pk 13.0 ● ● ● ● 30304372 Adapter, 2x 30ml D26mm RB, 2/pk 12.5 -1.6 ● ● ● ● 30304373 Adapter, 1x 50ml D29mm RB, 2/pk 12.0 -2.1 ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● 30304374 Adapter, 1x 50ml D30mm FA, 2/pk 11.7 -2.4 30559412 Adapter 15x 1.5/2.0ml D11mm MTBx 2 V1 9.5 13.6 11.8 Blank Adapter 1x 250ml D60mm CBCx 3 V1 Adapter 1x 175/225ml D61mm FAx 2 V1 NO Bucket/Adapter -4.6 -0.5 -0.5 30314842 Rack, 1x 100ml D46mm RB, 2/pk 14.1 30314843 Rack, 1x 85ml D38mm RB, 2/pk 14.1 30314844 Rack, 1x 50ml D34mm RB, 2/pk 30314845 30314846 30559414 Rotor, Sw ing out, 4x 100ml, ID Adapter, 5x 10ml D16mm RB, 2/pk -1.1 30559413 30314822 Correcció n (cm) FC5816R Rotor, Sw ing out, 4x 250ml, ID Radio (cm) FC5816 30314820 Descripció n del Contenedor/Adaptador FC5718R 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable Contenedo r/Adaptad or FC5718 Descripció n del rotor FC5714 Rotor 12.8 -2.3 14.8 ● ● ● -0.7 ● ● ● -0.7 ● ● ● 14.2 -0.6 ● ● ● Rack, 1x 50ml D29mm RB, 2/pk 14.2 -0.6 ● ● ● Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.7 ● ● ● 30314847 Rack, 1x 50ml D29mm Rim, 2/pk 14.3 -0.5 ● ● ● 30314848 Rack, 1x 30ml D25mm Rim, 2/pk 13.8 -1.0 ● ● ● 30314849 Rack, 1x 25ml D25mm RB, 2/pk 14.4 -0.4 ● ● ● 30314850 Rack, 2x 15ml D17mm FA, 2/pk 14.8 0.0 ● ● ● 30314851 Rack, 1x 15ml D17mm Rim, 2/pk 14.5 -0.3 ● ● ● ES-42 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314852 Rack, 7x 15ml D17mm RB, 2/pk 14.5 -0.3 ● ● ● 30314853 Rack, 9x 5-7ml D13mm RB, 2/pk 14.2 -0.6 ● ● ● 30314854 Rack, 10x 1.5ml D11mm, 2/pk 9.9 -4.9 ● ● ● 30314855 Rack, 9x 4.5ml D12mm FB, 2/pk 14.0 -0.8 ● ● ● 30314856 Rack, 7x 5-7ml D13mm RB, 2/pk 11.6 -3.2 ● ● ● 30314857 Rack, 7x 5-7ml D13mm Vac, 2/pk 14.2 -0.6 ● ● ● 30314858 Rack, 5x 4-10ml D16mm Sar, 2/pk 14.3 -0.5 ● ● ● Blank 14.6 ● ● ● 30314861 NO Bucket/Adapter Bucket, 100ml w o Cap, Sealable, 2/pk Bucket, 100ml w / Cap, Sealable, 2/pk 30314864 30314860 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 14.2 -0.4 ● ● ● 14.2 -0.4 ● ● ● Rack, 1x 100ml D40mm RB, 2/pk 13.8 -0.8 ● ● ● 30314865 Rack, 1x 85ml D38mm RB, 2/pk 14.0 -0.6 ● ● ● 30314866 Rack, 1x 50ml D34mm RB, 2/pk 14.0 -0.6 ● ● ● 30314867 Rack, 1x 50ml D29mm RB, 2/pk 14.0 -0.6 ● ● ● 30314868 Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.5 ● ● ● 30314869 Rack, 1x 50ml D29mm Rim, 2/pk 13.9 -0.7 ● ● ● 30314870 Rack, 1x 30ml D25mm Rim, 2/pk 13.1 -1.5 ● ● ● 30314871 Rack, 1x 25ml D25mm RB, 2/pk 14.0 -0.6 ● ● ● 30314872 Rack, 1x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314873 Rack, 2x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314874 Rack, 2x 15ml D17mm Rim, 2/pk 14.0 -0.6 ● ● ● 30314875 Rack, 4x 15ml D17mm RB, 2/pk 14.0 -0.6 ● ● ● 30314876 Rack, 7x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314877 Rack, 5x 1.5ml D11mm, 2/pk 14.0 -0.6 ● ● ● 30314878 Rack, 3x 10ml D17mm Sar, 2/pk 14.0 -0.6 ● ● ● 30314879 Rack, 4x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314880 14.0 -0.6 ● ● ● 14.6 0.0 ● ● ● 14.6 0.0 ● ● ● 30314883 Rack, 4x 10ml D17mm Vac, 2/pk Bucket, 2x 50ml D29mm FA, w / Rack, 2/pk Bucket, 3x 15ml D17mm FA, w / Rack, 2/pk Cy to, Rectangular Bucket, 2/pk 6.3 -8.3 ● ● ● Blank NO Bucket/Adapter 12.0 ● ● ● ● ● ● ● 30314891 Adapter, 48x 1.5/2.0ml D11mm, 2/pk 12.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 10.3 ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 30314894 30314895 30314881 30314882 30314824 30314825 30314826 30314827 Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID Rotor, Angle, 0.0 9.6 -0.7 ● ● Adapter, 1x 30ml D25mm RB, 2/pk 9.6 -0.7 ● ● Adapter, 1x 50ml D29mm RB, 2/pk 10.0 -0.3 ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 9.8 -0.5 ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 9.3 -1.0 ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 10.3 0.0 ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 9.5 -0.8 ● ● Blank NO Bucket/Adapter 11.5 ● ● ● ● ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 10.9 -0.6 ● ● ● ● ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 11.1 -0.4 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● Blank NO Bucket/Adapter ● ● ● ● ● ● 9.2 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4x 85ml RB, ID, Hi 30314828 30314829 30314830 30314831 30314832 30314834 30314835 30314836 30314838 30314839 Rotor, Sw ing out, 16x 50ml, ID Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable Rotor, Angle, 24x 1.5/2.0ml, ID Rotor, Angle, 12x 1.5/2.0ml, ID 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 8.5 -0.7 ● ● ● ● ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 8.4 -0.8 ● ● ● ● ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 8.9 -0.3 ● ● ● ● ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 8.6 -0.6 ● ● ● ● ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 8.3 -0.9 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 7.5 -1.7 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 8.3 -0.9 ● ● ● ● ● ● Blank NO Bucket/Adapter 16.5 ● ● 30304375 Rack, 4x 50ml D29mm FA, 2/pk 16.5 0.0 ● ● 30314583 Rack, 10x 15ml D17mm FA, 2/pk 16.5 0.0 ● ● Blank NO Bucket/Adapter 13.0 ● ● ● ● ● ● 30130894 Adapter, 1x 15ml D17mm RB, 2/pk 12.8 -0.2 ● ● ● ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 12.2 -0.8 ● ● ● ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 10.4 -2.6 ● ● ● ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 8.9 -4.1 ● ● ● ● ● ● Blank NO Bucket/Adapter 11.0 ● ● ● 30130891 Adapter, 1x 30ml D26mm, 2/pk 10.7 -0.3 ● ● ● 30130892 Adapter, 1x 16ml D18mm, 2/pk 10.3 -0.7 ● ● ● 30130893 Adapter, 1x 15ml D17mm, 2/pk 10.6 -0.4 ● ● ● 30130894 Adapter, 1x 15ml D17mm RB, 2/pk 10.6 -0.4 ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 10.2 -0.8 ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 8.3 -2.7 ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 6.7 -4.3 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130891 Adapter, 1x 30ml D26mm, 2/pk 8.2 -0.2 ● ● ● ● ● ● 30130892 Adapter, 1x 16ml D18mm, 2/pk 7.9 -0.5 ● ● ● ● ● ● 30130893 Adapter, 1x 15ml D17mm, 2/pk 8.0 -0.4 ● ● ● ● ● ● 30314892 Adapter, 1x 15ml D17.5mm FA, 2/pk 7.7 -0.7 ● ● ● ● ● ● Blank NO Bucket/Adapter 12.5 ● ● ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 12.2 -0.3 ● ● ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 10.5 -2.0 ● ● ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 9.0 -3.5 ● ● ● ● ● Blank NO Bucket/Adapter 11.0 ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 10.6 -0.4 ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 9.1 -1.9 ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 7.7 -3.4 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.2 ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.7 -0.7 ● ● ● ● ● ● Blank NO Bucket/Adapter 9.4 ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 9.1 -0.3 ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 8.4 -1.0 ● ● ● ● ● Blank NO Bucket/Adapter 8.5 ● ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.3 ● ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 -1.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 6.5 ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 6.3 -0.2 ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk Adapter, 1x 1.5ml for ov er16000x g, 6/pk NO Bucket/Adapter 5.6 -0.9 ● ● 6.4 -0.1 ● ● ● ● 30314900 30553085 Rotor Sw ingOut ES-43 Blank 19.3 ES-44 Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4x 750ml D99mm ID SL 30553138 Adapter 32x 7ml D13mm RBx 2 V1 18.7 -0.6 ● ● 30553135 Adapter 18x 15ml D17mm RBx 2 V1 18.8 -0.5 ● ● 30553132 Adapter 8x 30ml D26mm RBx 2 V1 19.1 -0.2 ● ● 30553131 Adapter 7x 50ml D29mm RBx 2 V1 19.1 -0.2 ● ● 30553127 Adapter 3x 100ml D41mm RBx 2 V1 19.1 -0.2 ● ● 30553125 18.9 -0.4 ● ● 19.1 -0.2 ● ● 30553128 Adapter 1x 250ml D62mm FBx 2 V1 Adapter 1x 225/190ml D61mm FAx 2 V1 Adapter 5x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553130 Adapter 7x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553133 Adapter 14x 15ml D17mm FAx 2 V1 19.2 -0.1 ● ● 30553134 Adapter 17x 15ml D17mm FAx 2 V1 Adapter 18x 5/10ml D16mm Vac/Sarx 2 V1 Adapter 38x 1.1/1.4ml D8mm Sarx 2 V1 Adapter 21x 10ml D16mm RBx 2 V1 19.0 -0.3 ● ● 19.0 -0.3 ● ● 18.9 -0.4 ● ● 18.9 -0.4 ● ● 18.8 -0.5 ● ● 30553126 30553140 30559377 30553136 30553139 Blank ● ● 13.7 -0.8 ● ● 13.4 -1.1 ● ● 12.4 -2.1 ● ● 12.6 -1.9 ● ● 30559421 Adapter 1x 250ml D61.5mm FBx 2 V1 Adapter 1x 175/225ml D61.5mm FAx 2 V1 Adapter 1x 50ml D29.5mm FAx 2 V1 Adapter 1x 250ml D60.5mm CBCx 2 V1 Adapter 4x 30ml D25.5mm RBx 2 V1 13.8 -0.7 ● ● 30559420 Adapter 3x 50ml D29mm RBx 2 V1 14.3 -0.2 ● ● 30559422 Adapter 6x 15ml D17mm FAx 2 V1 13.8 -0.7 ● ● 30564850 30559417 30553086 Rotor Angle 4x 500ml D69mm ID Adapter 27x 1.8/7ml D13mm Vac/Sarx 2 V1 NO Bucket/Adapter 30559419 30559416 14.5  Centrífuga FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ES-45 11.7 Tabla 7: Formulario de devolución / Certificado de descontaminación Apellido: _____________________________________ Organización / compañía: _____________________________________ Calle: _____________________________________ CÓDIGO POSTAL: _____________ lugar:____________________ Teléfono: _____________ fax:____________________ Correo electrónico: _____________________________________ Pos. Crowd Decontaminated object Serial number Llene en letras de molde mayúsculas Incluya este formulario con todas las devoluciones de equipo y ensambles. La declaración completa acerca de la descontaminación es un prerrequisito para la asunción y posterior procesamiento de la devolución. Si no se incluye la explicación correspondiente, llevamos a cabo la descontaminación con costos a su cargo. Description / Comment 1 2 3 4 ¿Hay partes de las mencionadas anteriormente en contacto con las siguientes sustancias? Soluciones, búferes, ácidos o álcalis acuosos que ponen en riesgo la salud:……… …. Sí No Agentes potencialmente infecciosos: ……………………………………………..………..  Sí No Reactivos y solventes orgánicos: …………………………………………………………...  Sí No Sustancias radioactivas: ……………………………………….………. α.. β.. γ… ..  Sí No Proteínas que ponen en riesgo la salud: …………………………………………..………. Sí No ADN: …………………………………………………………………………………..………..  Sí No ¿Estas sustancias han tocado el equipo o ensamble? …………………………………… Sí No ¿Cuál, en caso afirmativo?: ________________________________________________________________ Descripción de las medidas para la descontaminación de las partes mencionadas: Confirmo la descontaminación adecuada: Compañía/Departamento ____________ Lugar y fecha: ______________________________ Firma de la persona autorizada: ________________________________________ 3ODQFURTXLVGX)&)&)&)& 1 2 7 4 3 6 5 Figure. 1  &URTXLVGHODFDUWHGX)&5 1 2 7 3 4 6 5 Figure. 2 &DUWHFURTXLVGX)&5)&5 1 2 7 3 6 4 5 Figure. 3 1 Couvercle de la centrifugeuse 3 Affichage 5 Contacteur secteur 7 Déverrouillage d'urgence 2 Fenêtre du rotor 4 Étiquette de fonction 6 Connexion de l'alimentation Étiquette de fonction Étiquette de fonction pour les FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figure. 4 1 4 5 6 7 8 2 3 9 10 11 12 Figure. 5 1. Affichage ACL 3. 5. Démarrage de la centrifugation Sélection et modèle d'intensité de l'accélération/de la décélération Modèle de réglage de l'heure 7. 9. Centrifugation par rotation courte/rapide 11. Arrêt de la centrifugation/de la configuration 2. Modèle de configuration du programme 4. Sélection et modèle RPM/RCF 6. Modèle de configuration de la température (FC5515R seulement) 8. Informations de la configuration du stockage 10. Déverrouillage du couvercle 12. Réglage du bouton/du cadran : Changement du numéro Affichage ACL L'image suivante présente les éléments individuels de l'affichage ACL. Figure. 6 M2 M3 M1 M12 M15 M13 M4 M9 M11 M10 M5 M6 M7 A-1 M8 A-2 M14 A-3 A-4 Figure. 7 Affichage des champs: A-1 Affichage champ – « RPM/RCF » A-2 Affichage champ – « Acc/Déc » « Service » A-3 Affichage champ – « Heure/Prog » A-4 Affichage champ – « Temp » Messages/logos de l'affichage des champs M1 M4 M7 M10 M13 « fermer » « Rotor n° » « accél » « programme » «h m s» M2 « ouvrir » M5 « tr/min » M8 « décél » M11 « erreur » M14 « température » M3 « rotor » M6 « rcf » M9 « rayon » M12 « service » M15 « pré-refroidissement » Rotor n° Tableau FC5816R FC5916 FC5916R Rotor, Angle, 24x1.5/2.0ml, ID, Sealable Rotor, Swing out, 4x250ml, ID Rotor, Angle, 6x250ml FB, ID Rotor, Swing out, 4x100ml, ID Rotor, Swing out, 4x100ml, ID, Sealable Rotor, Swing out, 2x3MTP, ID Rotor, Angle, 6x85ml RB, ID, Hi Rotor, Angle, 6x85ml RB, ID Rotor, Angle, 4x85ml RB, ID, Hi Rotor, Swing out, 16x50ml, ID Rotor, Angle, 10x50ml FA, ID Rotor, Angle, 6x50ml RB/FA, ID Rotor, Angle, 6x50ml RB, ID, Hi Rotor, Angle, 30x15ml RB/FA, ID Rotor, Angle, 20x10ml RB, ID, Hi Rotor, Angle, 12x15ml RB/FA, ID Rotor, Angle, 44x1.5/2.0ml, ID Rotor, Angle, 30x1.5/2.0ml, ID, Sealable Rotor, Angle, 24x1.5/2.0ml, ID Rotor, Angle, 12x1.5/2.0ml, ID Rotor, Angle, 64x0.5ml, ID Rotor, Angle, 4x8-w PCR Strip, ID Rotor SwingOut 4x750ml D99mm ID SL Rotor Angle 4x500ml D69mm ID FC5816 30304361 30314820 30314821 30314822 30314823 30314824 30314825 30314826 30314827 30314828 30314829 30314830 30314831 30314832 30314833 30314834 30314835 30314836 30314838 30314839 30314840 30314841 30553085 30553086 Capacité FC5718R 61 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 38 39 40 41 85 86 Compatible FC5718 N° de command e FC5714 Affichage du numéro de rotor ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● TABLE DES MATIÈRES 1. 2. 3. 4. 5. 6. INTRODUCTION............................................................................................................................................................................ 1 1.1 DESCRIPTION ............................................................................................................................................................................... 1 1.2 CARACTÉRISTIQUES ........................................................................................................................................................................... 1 1.3 DEFINITION DES SYMBOLES ET DES INDICATEURS D'AVERTISSEMENTS ....................................................................................................... 1 1.4 PRECAUTIONS DE SECURITE ................................................................................................................................................................ 2 1.4.1 Utilisateur ........................................................................................................................................................................... 2 1.4.2 Rotor et accessoires........................................................................................................................................................... 2 1.4.3 Mesures pour votre protection ......................................................................................................................................... 2 1.4.4 Exclusions d'influences environnementales suivantes ................................................................................................... 3 1.4.5 Mesures pour un fonctionnement sans danger .............................................................................................................. 3 1.4.6 Dangers et précautions ..................................................................................................................................................... 3 INSTALLATION.............................................................................................................................................................................. 4 2.1 DEBALLAGE...................................................................................................................................................................................... 4 2.1.1 Emballage de livraison ...................................................................................................................................................... 4 2.2 SÉLECTION DE L'EMPLACEMENT .......................................................................................................................................................... 4 2.3 I NSTALLATION .................................................................................................................................................................................. 4 2.4 PRECAUTIONS DE SECURITE PENDANT LE FONCTIONNEMENT ET GARANTIE ................................................................................................ 5 FONCTIONNEMENT ..................................................................................................................................................................... 5 3.1 MONTAGE ET CHARGEMENT DU ROTOR ............................................................................................................................................... 5 3.1.1 Installation des rotors........................................................................................................................................................ 5 3.1.2 Chargement des rotors angulaires................................................................................................................................... 6 3.1.3 Chargement des rotors horizontaux ................................................................................................................................ 7 3.1.4 Chargement et surcharge des rotors ............................................................................................................................... 7 3.1.5 Démontage du rotor .......................................................................................................................................................... 7 3.2 CONTROLE DU COUVERCLE ................................................................................................................................................................. 7 3.2.1 Ouverture du couvercle ..................................................................................................................................................... 7 3.2.2 Verrouillage du couvercle ................................................................................................................................................. 8 3.3 PRÉSÉLECTION.................................................................................................................................................................................. 8 3.3.2 Présélection des durées de centrifugation ...................................................................................................................... 9 3.3.3 Présélection de l'intensité de freinage et de l'accélération........................................................................................... 9 3.3.4 Présélection de la température (Seulement les modèles réfrigérés).......................................................................... 10 3.3.5 Pré-refroidissement (Seulement les modèles réfrigérés)............................................................................................. 10 3.4 CORRECTION DU RAYON .................................................................................................................................................................. 11 3.5 PROGRAMMATION.......................................................................................................................................................................... 11 3.5.1 Mémorisation des programmes ..................................................................................................................................... 11 3.5.2 Rappel des programmes mémorisés.............................................................................................................................. 12 3.5.3 Sortie du mode de programmation................................................................................................................................ 12 3.6 DEMARRAGE ET ARRET DE LA CENTRIFUGEUSE .................................................................................................................................... 12 3.6.1 Démarrage de la centrifugeuse...................................................................................................................................... 12 3.6.2 La touche « STOP » .......................................................................................................................................................... 13 3.7 DETECTION D'UN DESEQUILIBRAGE.................................................................................................................................................... 13 PARAMÉTRAGE.......................................................................................................................................................................... 14 4.1 RÉGLAGES DE BASE ......................................................................................................................................................................... 14 4.1.1 Accès au mode « Données de fonctionnement ».......................................................................................................... 14 4.1.2 Indication de la température.......................................................................................................................................... 14 4.1.3 Signal Marche/Arrêt ........................................................................................................................................................ 15 4.1.4 Volume de présélection du signal sonore...................................................................................................................... 15 4.1.5 Sélection de la sonorité du signal sonore de fin de centrifugation............................................................................. 16 4.1.6 Marche/Arrêt du son du clavier ..................................................................................................................................... 16 4.1.7 Appel des données de fonctionnement.......................................................................................................................... 16 MAINTENANCE........................................................................................................................................................................... 17 5.1 MAINTENANCE ET NETTOYAGE ......................................................................................................................................................... 17 5.1.1 Entretien général : ........................................................................................................................................................... 17 5.1.2 Nettoyage et désinfection de l'unité.............................................................................................................................. 18 5.1.3 Nettoyage et désinfection du rotor................................................................................................................................ 18 5.1.4 Désinfection des rotors en aluminium........................................................................................................................... 18 5.1.5 Désinfection des rotors PP .............................................................................................................................................. 18 5.1.6 Bris de verre...................................................................................................................................................................... 19 DÉPANNAGE............................................................................................................................................................................... 19 6.1 MESSAGE D'ERREUR : CAUSES /SOLUTIONS......................................................................................................................................... 19 6.2 ÉTUDE DES MESSAGES D'ERREUR POSSIBLES ET LEURS SOLUTIONS .......................................................................................................... 19 6.2.1 Libération du couvercle pendant une panne d'alimentation (libération en cas d'urgence) .................................... 19 6.2.2 Description du système de messages d'erreur.............................................................................................................. 20 7. RÉCEPTION DES CENTRIFUGEUSES POUR RÉPARATION........................................................................................................ 21 8. TRANSPORT, STOCKAGE ET MISE AU REBUT.......................................................................................................................... 21 8.1 TRANSPORT ................................................................................................................................................................................... 21 8.2 STOCKAGE ..................................................................................................................................................................................... 21 9. DONNEES TECHNIQUES............................................................................................................................................................. 23 9.1 SPECIFICATIONS.............................................................................................................................................................................. 23 9.1.1 Centrifugeuse FC5714 ...................................................................................................................................................... 23 9.1.2 Centrifugeuse FC5718 ...................................................................................................................................................... 24 9.1.3 Centrifugeuse FC5718R.................................................................................................................................................... 25 9.1.4 Centrifugeuse FC5816 ...................................................................................................................................................... 26 9.1.5 Centrifugeuse FC5816R.................................................................................................................................................... 27 9.1.6 Centrifuge FC5916 ............................................................................................................................................................ 28 9.1.7 Centrifuge FC5916R.......................................................................................................................................................... 29 9.2 SCHEMAS ET DIMENSIONS................................................................................................................................................................ 30 10. CONFORMITÉ ........................................................................................................................................................................ 31 11. ANNEXE.................................................................................................................................................................................. 32 11.1 TABLE 1: POIDS NET AUTORISE ................................................................................................................................................ 33 11.2 TABLE 2: TEMPERATURES LES PLUS FAIBLES A LA VITESSE MAXIMUM .......................................................................... 35 11.3 TABLE 3: VITESSE ET VALEURS RCF MAXI POUR LES ROTORS AUTORISES ...................................................................... 35 11.4 TABLE 4: DUREES D'ACCELERATION ET DE DECELERATION.................................................................................................... 39 11.5 TABLEAU 5 : MESSAGES D'ERREUR ........................................................................................................................................... 41 11.6 TABLE 6: CORRECTION DU RAYON........................................................................................................................................... 42 11.7 TABLEAU 7 : FORMULAIRE DE RACHAT/CERTIFICAT DE DECONTAMINATION .......................................................................................... 46 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-1 1. INTRODUCTION 1.1 DESCRIPTION Merci d'avoir choisi ce produit OHAUS. Tous les symboles indiquent des instructions de sécurité et des emplacements présentant des situations potentiellement dangereuses. Veuillez lire le manuel complètement avant d'utiliser les centrifugeuses Frontier TM MultiPro pour éviter un fonctionnement incorrect. Les centrifugeuses FrontierTM MultiPro ont été conçues pour la séparation de matériaux ou de mélanges de différentes densités. Les centrifugeuses OHAUS sont conçues pour être utilisées exclusivement en intérieur et par du personnel qualifié. 1.2 Caractéristiques Les modèles FC5714/FC5718/FC5816/FC5916 sont des centrifugeuses universelles non-réfrigérées. Les modèles FC5718R/FC5816R/FC5916R sont des centrifugeuses universelles réfrigérées. Les trois modèles sont proposés avec deux tensions d’alimentation, 230 V ou 120 V Les centrifugeuses sont utilisées avec des rotors horizontaux ou angulaires Les réglages sont accessibles via des boutons et sélectionnés via le régleur central. Toutes les valeurs préréglées et les valeurs courantes apparaissent en continu sur l’écran ACL. La centrifugeuse est entraînée par un moteur asynchrone sans entretien Les données techniques détaillés sont consultables dans «Données techniques» 1.3 Définition des symboles et des indicateurs d'avertissements Les remarques de sécurité sont signalées par des mots indicateurs et par des symboles d'avertissement. Elles présentent les problèmes et les avertissements de sécurité. L'inobservation des remarques de sécurité peut conduire à des accidents, à l'endommagement de l'instrument, à des dysfonctionnements et à des résultats erronés. Le niveau du danger fait partie d'une note de sécurité et réalise la distinction entre les résultats possibles d'une inobservation pour chacun d'entre eux. Mots indicateurs DANGER AVERTISSEMENT ATTENTION ATTENTION NOTE Possibilité d'un accident grave, voire mortel, s'il n'est pas évité. Pour une situation dangereuse avec un risque moyen pouvant être à l'origine d'un accident ou d'un décès, si elle n'est pas évitée. Pour une situation dangereuse avec un risque faible, pouvant être à l'origine d'un endommagement de l'appareil ou des biens, de perte des données, ou de blessures si elle n'est pas évitée. Pour des informations importantes sur le produit. Peut conduire à des détériorations de l'équipement si elle n'est pas évitée Pour des informations utiles sur le produit Symboles d'avertissement Danger général Danger de choc électrique Tension alternative Risques biologiques Explosion Écrasement FR-2 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Signalisation informative et d'avertissement sur la surface de la centrifugeuse Quatre supports doivent être utilisés à tout moment sur les quatre emplacements de rotation des rotors, sinon des détériorations de la centrifugeuse se produiront. De telles détériorations ne seront pas couvertes par la garantie de ce produit. Attention ! Vérifier le serrage de l'écrou du rotor avant chaque utilisation. Retirer le bouchon secteur avant d'ouvrir l'enceinte ou d'effectuer un déverrouillage d'urgence. Sens de la rotation, rotation horaire pour l’entraînement du rotor Référence pour le chargement des rotors 1.4 Précautions de sécurité 1.4.1 Utilisateur Les centrifugeuses Ohaus sont exclusivement destinées à une utilisation intérieure et par un personnel qualifié. Cet appareil ne peut être utilisé que par un personnel dûment formé. Ce personnel devra lire intégralement le manuel de fonctionnement et se familiariser au fonctionnement de l'appareil. 1.4.2 Rotor et accessoires Seuls des rotors et des accessoires d'origine d'OHAUS seront utilisés. Toute autre utilisation ou utilisation envisagée est considérée comme inadéquate. OHAUS n'est pas responsable des détériorations résultant d'une utilisation inadéquate. ATTENTION : Lire tous les avertissements de sécurité avant d’effectuer l’installation, les connexions ou la maintenance de cet équipement. Si ces avertissements ne sont pas respectés, des accidents et/ou des dommages aux biens peuvent en résulter. Conserver ces instructions pour référence ultérieure. 1.4.3 Mesures pour votre protection AVERTISSEMENT : Ne jamais travailler dans un environnement pouvant présenter des dangers d'explosion ! L'enceinte de l'instrument n'est pas étanche aux gaz. (Danger d'explosion en prés ence d'étincelles, corrosion provoquée par l'introduction de gaz) AVERTISSEMENT : Lors de l'utilisation de produits chimiques et de solvants, se conformer aux instructions du fabricant et aux règles générales de sécurité du laboratoire. AVERTISSEMENT :La centrifugeuse n'est pas étanche. Des mesures de protection adaptées seront prises lors de l'utilisation de la centrifugeuse pour des échantillons infectieux ou pathogènes. Des précautions de sécurité appropriées doivent être appliquées lors de la manipulation de ces échantillons. Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-3 1.4.4 Exclusions d'influences environnementales suivantes       Vibrations puissantes Lumière directe du soleil Humidité atmosphérique supérieure à 80 % Présence de gaz corrosifs Températures supérieures à 35 °C et inférieures à 2 °C Champs électriques ou magnétiques puissants : AVERTISSEMENT : Risque de chocs électriques dans le boîtier. Le boîtier doit être ouvert uniquement par un personnel qualifié et autorisé. Enlever toutes les connexions d’alimentation à l’unité avant l'ouverture. 1.4.5 Mesures pour un fonctionnement sans danger  Ne jamais dévisser les deux moitiés de l'enceinte  Sécher immédiatement tout déversement de liquide. L'instrument n'est pas étanche  Vérifier que la plage de tension d'entrée et le type de prise de l'équipement sont compatibles avec la tension secteur locale.  Brancher le câble d'alimentation uniquement à une prise d'alimentation électrique correctement mise à la masse.  Utiliser uniquement un câble d'alimentation avec une puissance nominale supérieure aux spécifications sur l'étiquette de l'équipement.  Positionner l'équipement pour qu'il soit facile de déconnecter le câble d'alimentation de la prise d'alimentation électrique.  S'assurer que le cordon d'alimentation n’est pas un obstacle potentiel ou ne présente aucun danger de chute.  Cet équipement est exclusivement destiné à un usage en intérieur. Utiliser l'équipement seulement dans des lieux secs.  Utiliser uniquement des accessoires approuvés.  Faire fonctionner l'équipement uniquement dans les conditions ambiantes spécifiées dans ces instructions.  Déconnecter l'équipement de l'alimentation électrique pour le nettoyer.  Ne pas utiliser l'équipement dans des environnements dangereux ou instables.  La maintenance doit être exclusivement exécutée par un personnel agréé. 1.4.6 Dangers et précautions Les précautions suivantes doivent être observées afin de protéger les personnes et l'environnement :  Pendant la centrifugation et la mise en œuvre de matières dangereuses, la présence de personnes est interdite dans un rayon de 30 cm autour de la centrifugeuse conformément aux réglementations EN 61010-2-020.  Les FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R ne sont pas protégées contre les explosions et ne doivent pas être utilisées dans des lieux ou dans des zones susceptibles aux explosions. La centrifugation de substances inflammables, explosives ou radioactives qui réagissent chimiquement en produisant une énergie élevée est strictement interdite. La décision finale quant aux risques associés à l'utilisation de telles substances incombe à l'utilisateur de la centrifugeuse.  Ne pas centrifuger de matières toxiques ou pathogènes sans précautions de sécurité adéquates, c'est-à-dire la centrifugation de godets/de tubes présentant une étanchéité peu hermétique ou défectueuse, ce qui est strictement interdit. L'utilisateur a l'obligation d'appliquer des procédures appropriées de désinfection si des substances dangereuses ont contaminé la centrifugeuse et/ou ses accessoires. Lors de la centrifugation de substances infectieuses, l'utilisateur prêtera attention aux précautions générales de laboratoire. Si nécessaire, contacter le responsable de la sécurité !  Il est interdit de faire fonctionner la centrifugeuse avec d'autres rotors que ceux répertoriés pour cette unité.  Le couvercle de la centrifugeuse ne doit jamais être ouvert tant que le rotor est en fonctionnement ou en rotation à une vitesse supérieure à 2 m/s FR-4 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 1.4.7 Abréviations utilisées dans ce manuel Symbol/Abbreviations Unit RPM [min-1] rpm RCF [x g] PCR Description revolutions per minute relative centrifugal force PCR Polymerase chain reaction PP - Polypropylen PC - Polycarbonat accel - acceleration decel - deceleration prog - program 2. INSTALLATION 2.1 Déballage Sortir avec précaution la centrifugeuse et chacun de ses composants hors de l'emballage. Les composants inclus varient en fonction du modèle de centrifugeuse (voir le tableau ci-dessous). Conserver l'emballage afin d'assurer un stockage et un transport en toute sécurité. Le manuel d'instructions doit toujours se trouver à proximité de la centrifugeuse ! Les rotors/les accessoires seront emballés séparément. AVERTISSEMENT : Risques associés au levage. Risque de blessures si une seule personne soulève l'équipement. Se faire aider lorsque l'équipement est soulevé ou déplacé. On préconise qu’il y ait au moins quatre personnes, voire davantage, pour déplacer et installer ces trois unités ou d’utiliser un matériel de transport adapté. Consulter la section 8.3 pour davantage de précisions sur la techniques de lavage pour la sortie de l’emballage. 2.1.1 Emballage de livraison Quantité 1 1 1 1 1 Description Centrifugeuse FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Câble d’alimentation Carte de garantie Manuel d’instructions / Guide de démarrage rapide Clé du rotor 2.2 Sélection de l'emplacement REMARQUE ! Éviter les vibrations excessives, les sources de chaleur, les courants d'air ou les variations rapides de température.  La centrifugeuse doit être installée sur une surface plane, de niveau et résistante, si possible dans une armoire/sur une table de laboratoire ou sur une surface solide sans vibrations.  Pendant la centrifugation, la centrifugeuse doit être placée de manière à présenter un espace minimum de 30 cm de chaque côté de l'unité conformément aux normes EN 61010-2-020.  La centrifugeuse ne doit pas être positionnée à proximité d'une fenêtre ou d'un chauffage, où elle pourrait être exposée à une chaleur excessive, en effet les performances de l'unité se fondent sur une température ambiante de 23 °C. 2.3 Installation Suivre ces étapes :  Vérifier que l'alimentation électrique correspond à celle spécifiée sur l'étiquette signalétique du fabricant, située sur le panneau arrière. Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-5  Pour les modèles FC5714/FC5718/FC5816, la ligne d'alimentation doit être protégée par un disjoncteur 10 A cote de circuit (type K).  Pour les modèles FC5718R/FC5816R /FC5916/FC5916R, la ligne d'alimentation doit être protégée par un disjoncteur 16 A cote de circuit (type K)  En cas d'urgence, un commutateur de sûreté doit être installé en dehors de la pièce afin de déconnecter l'alimentation sur l'unité.  La centrifugeuse doit être connectée à une prise secteur reliée à la terre.  Raccorder la centrifugeuse au réseau. (La prise destinée au câble d’alimentation doit être d’un accès facile, de même que facile à débrancher.  Mettre l'instrument sous tension avec le commutateur d'alimentation.  Ouvrir le couvercle en utilisant le bouton Door Open (porte ouverte).  Retirer le dispositif sécurisant le mécanisme d'entraînement du moteur. 2.4 Précautions de sécurité pendant le fonctionnement et garantie  Ne pas faire fonctionner la centrifugeuse si elle n'est pas correctement installée.  Ne pas s'appuyer contre la centrifugeuse pendant le fonctionnement.  Ne pas rester dans la zone de sécurité de 30 cm autour de l'équipement plus longtemps que nécessaire pour des raisons de fonctionnement.  Ne laisser aucun produit potentiellement dangereux dans la zone de sécurité de 30 cm autour de l'équipement.  Ne pas faire fonctionner la centrifugeuse lorsqu'elle est démontée (sans son enceinte).  Ne pas faire fonctionner la centrifugeuse lorsque des composants mécaniques ou électriques ont subi des altérations.  Ne pas utiliser d'accessoires tels que des rotors ou des godets qui ne sont pas exclusivement approuvés par OHAUS Corporation, hormis les tubes pour centrifugeuse disponibles commercialement et fabriqués en verre ou en plastique.  Ne pas centrifuger de substances fortement corrosives qui pourraient endommager ou fragiliser les matériels.  Ne pas faire fonctionner la centrifugeuse avec des rotors ou des godets présentant des signes de corrosion ou de détériorations mécaniques.  Le fabricant est responsable de la sécurité et de la fiabilité de la centrifugeuse, seulement si :  L'unité est exploitée conformément à ce manuel d'instructions.  Des modifications, des réparations ou d'autres réglages sont réalisés par un personnel agréé et l'installation électrique est conforme à la réglementation en la matière. REMARQUE! Garantie La centrifugeuse a fait l’objet d’une vérification complète et de contrôles de qualité. Dans l’éventualité, peu probable, de vices de fabrication, la centrifugeuse et les rotors sont couverts par la garantie. Ladite garantie est nulle en cas de mauvaise utilisation, de dommages et de négligence, également dans le cas de l’utilisation de pièces de rechange et d’accessoires inadaptés, voire de la modification non-autorisée de la machine. Le droit d’apporter des modifications relève de la seule responsabilité du fabricant en vue d’apporter des améliorations! 3. FONCTIONNEMENT 3.1 Montage et chargement du rotor 3.1.1 Installation des rotors Nettoyer l'arbre d'entraînement ainsi que le collier de serrage avec un chiffon propre et sans graisse. Placer le rotor sur l'arbre d'entraînement. (Voir figure ci-dessous). Ne pas oublier que le rotor est intégralement installé sur l'arbre du moteur. FR-6 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figure. 8 Écrou du rotor Outil du rotor avec écrou Outil du rotor sans écrou Figure. 9 Couvercle à enclenchement Couvercle à vissage Figure. 10 Maintenir le rotor d'une main et fixer celui-ci sur l'arbre en tournant la vis de fixation dans le sens horaire. Serrer l'écrou de fixation avec la clé pour rotor fournie (voir les figures 9 et 10). Nous fournirons un outil pour les rotors sans écrou avec la centrifugeuse, l'outil pour le rotor avec écrou sera fourni en même temps que le rotor. ATTENTION : Vérifier que la vis de fixation est correctement installée avant chaque centrifugation. (Se reporter à la figure 11) Ne pas faire fonctionner la centrifugeuse avec des rotors ou des godets présentant des signes de corrosion ou de détériorations mécaniques. Ne pas faire fonctionner la centrifugeuse avec des substances fortement corrosives, ce qui pourrait endommager le rotor, les godets et les matériels. En cas de questions, contacter le fabricant ! 3.1.2 Chargement des rotors angulaires Les rotors doivent être chargés symétriquement et avec un poids égal (voir la figure ci-dessous). L'adaptateur ne peut être chargé qu'avec les récipients appropriés. La différence de poids entre les récipients remplis doit être autant que faire se peut réduite au minimum. C'est pourquoi nous recommandons leur pesée avec une balance. Ceci réduit l'usure de l'entraînement et le bruit de centrifugation. La charge maximum par orifice est définie sur chaque rotor. Figure. 11 INCORRECT (6 tubes) Figure. 12 CORRECT Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-7 3.1.3 Chargement des rotors horizontaux Le chargement des godets ou des cuves doit être réalisé conformément à la figure ci-dessous. On peut exploiter, par exemple, un rotor à quatre logements avec seulement deux godets. Néanmoins, les godets chargés doivent être en vis-à-vis. S’assurer que les godets non chargés soient également placés à l’intérieur du rotor (cf. ci-dessous).En pr incipe, il n’est pas per mis de mettre en œuvre les rotors jusqu’à ce que tous les godets ou râteliers soient installés dans le rotor. Les boulons de rotor doivent être lubrifiés avec du «High TEF Oil». Les tubes à essai doivent être remplis visuellement de façon uniforme et disposés dans les orifices ou les râteliers à tubes. La différence en poids des godets chargés de doit pas dépasser environ 1,0 g. ATTENTION!’ Il n’est per mis de mettre en œuvre les rotors horizontaux uniquement que si tous les emplacements sont occupés, soit avec quatre godets, soit avec quatre porteurs; ne pas mélanger godets et porteurs !!. ATTENTION! Ne pas mettre en œuvre la centrifugeuse avec des rotors ou des godets présentant des signes de corrosion ou de fatigue mécanique. Ne pas mettre en œuvre avec des substances extrêmement corrosives qui pourraient endommager le rotor ou des godets. Contacter le fabricant pour toute question ! Figure.13 WRONG Figure.14 CORRECT 3.1.4 Chargement et surcharge des rotors Tous les rotors approuvés sont répertoriés avec leur vitesse maximum ainsi que leur poids de remplissage maximum sur le « tableau 2 du poids net autorisé » (voir ANNEXE). La charge maximum autorisée pour un rotor, déterminée par le fabricant, ainsi que sa vitesse maximum autorisée (consulter l'étiquette sur le rotor), ne doivent pas être dépassées. Les liquides chargés sur les rotors doivent présenter une densité homogène maximum de 1,2 g/ml ou moins lorsque le rotor fonctionne à sa vitesse maximum. Pour centrifuger des liquides d'une densité plus élevée, la vitesse doit être réduite conformément à la formule suivante : 1,2 Vitesse réduite nrouge = √ 𝑑𝑒𝑛𝑠𝑖𝑡é 𝑝𝑙𝑢𝑠 é𝑙𝑒𝑣é𝑒 x vitesse maxi (nmaxi) du rotor Exemple : nrrouge = √ 1,2 1,7 x 4 000 = 3 360 tr/min En cas de questions, contacter le fabricant SVP ! Afin de déterminer la force centrifuge relative (FCR/Force g) pour un adaptateur particulier, on peut calculer en DIN 58 970 à l’aide de la formule jointe: FCR = 1.117862 *10-5*n2*rmax n: tours par minute (tr/min.) rmax: rayon max. de centrifugation en cm à la base des tubes. 3.1.5 Démontage du rotor Desserrer complètement l'écrou de fixation du rotor (dévisser au-delà du point de résistance) et soulever le rotor verticalement hors de la centrifugeuse. (Voir les figures 10 et 11) 3.2 Contrôle du couvercle 3.2.1 Ouverture du couvercle Après une centrifugation, lors de la fermeture du couvercle de la centrifugeuse, le mot "close" (M1) apparaît sur l'affichage "RPM | RCF" (A-1). De plus, si un rotor se trouve dans la centrifugeuse, le mot "rotor"(M3) apparaît Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-8 ainsi que le numéro de code du rotor s'y trouvant installé "71" (M4). En l'absence de rotor dans la centrifugeuse, le mot "rotor"(M3) clignote et apparaît en supplément le mot "no" (M4). En appuyant sur la touche "Door Open" (7), l'utilisateur pourra libérer le couvercle de la centrifugeuse. Aussitôt que le couvercle électromagnétique est complètement libéré, le mot "open" (M2) apparaît. Il est dorénavant possible d'ouvrir le couvercle de la centrifugeuse. Se reporter à la figure 15 ci-dessous pour référence. Figure. 15 Pendant la centrifugation, l'utilisateur peut rechercher le type de rotor à tout moment en appuyant sur la touche "Door Open" (7). 3.2.2 Verrouillage du couvercle Le couvercle ne doit être que très légèrement abaissé. Un verrouillage électromagnétique du couvercle le referme et simultanément, le mot "open" (M2) disparaît (voir la figure 15). Afin de signaler que la centrifugeuse est sur le point de démarrer, le mot "close" (M1) apparaît sur l'affichage "RPM | RCF" (A-1). Simultanément le mot "rotor" (M3) s'affiche ainsi que le n° de code du rotor installé dans la centrifugeuse, "no 71" (M4). Ainsi, toutes les données spécifiques du rotor, telles que la vitesse max imum, l'accélération, etc., sont adoptées. ATTENTION : Ne pas se coincer les doigts entre le couvercle et l'appareil ou le mécanisme de verrouillage lors de la fermeture du couvercle ! 3.3 Présélection 3.3.1 Présélection de la vitesse/de la valeur RCF Cette présélection est activée au moyen de la touche "RPM | RCF" (4) (voir la figure 16 ci-dessous). En appuyant sur la touche une fois, le mot "rpm" (M5) clignote. En appuyant sur la touche deux fois, il est possible de présélectionner les forces centrifuges. Le mot clignotant "rcf" (M6) apparaît alors. Il est possible de définir les valeurs désirées avec le bouton de réglage (1). Sur l'affichage (A-1), la valeur régulée est présentée en permanence, avant, pendant et après la centrifugation. Figure. 16 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-9 Tant qu'aucun rotor n'est inséré, la vitesse est réglable entre 200 tr/min et le nombre maximum de révolutions de la centrifugeuse. Si un rotor se trouve dans la centrifugeuse, la vitesse ne peut être présélectionnée que jusqu'au maximum autorisé de révolutions de ce rotor. C'est la même que la présélection de la valeur RCF. La plage de réglage se situe entre la force centrifuge à 200 tr / min et la force centrifuge maximale admissible du rotor. Voir le «Tableau 3 : vitesse maximum et valeurs RCF des rotors utilisés » (voir ANNEXE). Toutes les valeurs importantes sont répertoriées ici. ATTENTION : L'utilisateur doit vérifier la rotation maximum autorisée des tubes à essai avec le fabricant. 3.3.2 Présélection des durées de centrifugation La durée de centrifugation peut être présélectionnée selon trois différentes plages de 10 secondes à 99 heures 59 minutes. 1. Plage de 10 secondes jusqu'à 59 minutes 50 secondes par étapes de 10 secondes 2. Plage d'une heure jusqu'à 99 heures 59 minutes par étapes d'une minute 3. Le fonctionnement en continu "cont" peut être interrompu avec la touche "Stop" (10) (voir la figure 17). La durée de fonctionnement peut être présélectionnée avec le couvercle ouvert ou fermé. Pour activer le réglage de la durée de fonctionnement, appuyez sur la touche "Time" (6). Sur l'affichage « Time/Prog" (A-3), l'indication "m : s" ou "h : m" clignote en fonction des réglages précédents. Pour définir la valeur désirée, utiliser le bouton de réglage (1). Après avoir dépassé 59 min 50 s, l'indication passe automatiquement sur "h : m". Après avoir dépassé 99 heures 59 min, le mot "cont" apparaît sur l'affichage "Time/Prog" (A-3). Ce fonctionnement en continu ne peut être interrompu qu'en appuyant sur la touche "Stop" (10). Le décompte temporel commence aussitôt que la vitesse prescrite est atteinte. L'affichage présente toujours la durée de fonctionnement restante. (Voir la figure 17) Figure. 17 3.3.3 Présélection de l'intensité de freinage et de l'accélération Cette fonction est activée avec la touche "Accel/Decel" (5) (voir la figure 18). En appuyant une fois sur la touche, le mot "accel" (M7) clignote sur l'affichage "Acc/Dec" (A-2). L'accélération désirée peut être présélectionnée en réglant le bouton (1). La valeur 0 est équivalente à la plus faible accélération et la valeur 9 correspond à la plus élevée. En appuyant deux fois sur la touche "Accel/Decel" (5), le mot "decel"(M8) apparaît sur l'affichage "Acc/Dec" (A2). Dorénavant, l'intensité de freinage désirée peut être présélectionnée en réglant le bouton (1). La valeur 9 est équivalente à la durée de freinage la plus courte et la valeur 0 correspond à la plus longue. Voir le «Tableau 5 : Durées d'accélération et de décélération » (ANNEXE). Les durées d'accélération et de décélération pour les étapes 0 à 9 d'accélération et de décélération concernant les rotors utilisés sont présentées ici. FR-10 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figure. 18 3.3.4 Présélection de la température (Seulement les modèles réfrigérés) Cette fonction est activée par la touche "Temp/Setup" (13). Après avoir appuyé sur cette touche de l'affichage "Time/Prog", l'indication "°C" (A-4) clignote. Avec le bouton de réglage (1), la température de test désirée peut être présélectionnée par incréments de 1 °C dans la plage de -20 °C à +40 °C. La valeur est indiquée en permanence sur l'affichage (figure 19), avant, pendant et après la centrifugation. Prendre note des températures les plus basses des rotors à la vitesse maximum ! Figure. 19 3.3.5 Pré-refroidissement (Seulement les modèles réfrigérés) Si les échantillons sont sensibles à la température, il est utile de pré-refroidir la centrifugeuse, le rotor et finalement les godets à la température requise de fonctionnement. Pour ce faire, insérer le rotor retenu et prédéfinir la température correspondante. En appuyant simultanément sur les touches "Temp/Setup" (13) (voir la figure 20) et "Time" (6), l'utilisateur démarrera une centrifugation. Pendant le fonctionnement, l'unité choisit automatiquement une vitesse de rotation équivalente à 20 % de celle autorisée du rotor inséré. Après avoir atteint la température prédéfinie, il est possible de quitter le pré-refroidissement avec la touche "Stop" (10). En fonction du rotor inséré, le pré-refroidissement dure environ de 10 à 20 minutes. Figure. 20 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-11 3.4 Correction du rayon L'utilisation d'adaptateurs ou de réducteurs peut modifier le rayon de centrifugation de chaque rotor. Dans ce cas, il est possible de modifier manuellement le rayon. Poursuivre de la manière suivante : Fermer le couvercle et maintenir ensuite appuyées les touches "Time" (6) (voir la figure 21) et "Prog/Setup" (11). Sur l'affichage "Time/Prog" (A-3), le mot "radius" (M9) apparaît. Avec le bouton de réglage (1), l'utilisateur peut présélectionner des corrections du rayon (voir Tableau 7, ANNEXE) par incrément de 0,1 centimètre. Après avoir défini une correction du rayon, le mot "radius" (M9) apparaît. Ce mot sera visible jusqu'à ce que la correction du rayon soit revenue à 0. Figure. 21 3.5 Programmation 3.5.1 Mémorisation des programmes Il est possible de mémoriser jusqu'à 99 centrifugations incluant tous les paramètres pertinents, notamment les rotors utilisés. Il est possible d'utiliser un numéro de programme libre et de le rappeler à nouveau. Installer le rotor nécessaire dans la centrifugeuse. En appuyant sur la touche "Prog/Setup" (11) de l'affichage "Time/Prog" (A-3), le mot "programm" apparaît. Avec le bouton de réglage (1), choisir le numéro de programme retenu. Si un numéro de programme est déjà occupé, les mots "rotor" (M3) et "xx" (M4) apparaîtront sur l'affichage "RPM | RCF" (A-1). Pour les numéros de programme libres, 0 apparaît. Figure. 22 Fermer le couvercle de la centrifugeuse. Procéder dorénavant selon la description précédente afin de définir tous les paramètres importants de la centrifugation. Si le couvercle n'est pas fermé lors de la mémorisation du programme, les mots "FirSt" et "CLOSE Lid" (voir la figure 23) clignotent alternativement sur l'affichage "RPM/RCF" (A-1). Si la centrifugation doit être démarrée sans mémorisation du programme, les mots "First" et "PrESS StoreE" (voir la figure 24) clignotent alternativement sur l'affichage "RPM/RCF" (A-1). FR-12 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figure. 23 Figure. 24 Pour adapter les données, appuyer sur la touche "Store" (12) (voir les figures 23 et 24) pendant 1 seconde environ. Si le programme est correctement mémorisé, le mot "StorE" apparaît sur l'affichage "RPM/RCF" (A-1). Ceci aura pour effet de faire disparaître le mot "programm" (M10). Aussitôt après avoir libéré la touche "Store" (12), le mot "programm xx" (M10) réapparaît avec xx correspondant à l'emplacement du programme choisi. Si tous les numéros de programme sont occupés, l'utilisateur peut choisir un ancien numéro qui n'est plus actuellement utile et y introduire simplement les nouveaux paramètres. 3.5.2 Rappel des programmes mémorisés Pour rappeler des programmes mémorisés, appuyer sur la touche "Prog/Setup" (11) (voir la figure 25) alors que le couvercle est déjà fermé. Sur l'affichage "Time/Prog" (A-3), le mot "programm –"(M10) apparaît. Le numéro de programme désiré peut être présélectionné avec le bouton de réglage (1). Sur les affichages respectifs, les valeurs mémorisées de ce programme apparaîtront. Si le rotor dans la centrifugeuse n'est pas celui correspondant au programme présélectionné, le mot "rotor" (M3) clignotera sur l'affichage "RPM | RCF" (A-1). Simultanément, le mot "FALSE" et le numéro du rotor mémorisé " xx" (M4) clignoteront alternativement. Figure. 25 3.5.3 Sortie du mode de programmation Pour quitter le mode de programmation, appuyer simplement sur la touche "Prog/Setup" (11) (voir la figure 25). Sur l'affichage "Time/Prog », le mot "programm" apparaîtra. Régler l'affichage sur "programm--" (M10) avec le bouton de réglage (1). 3.6 Démarrage et arrêt de la centrifugeuse 3.6.1 Démarrage de la centrifugeuse La centrifugeuse peut être démarrée avec la touche "Start" (9) (voir la figure 26) ou avec la touche "Quick Spin" (8).Avec la touche "Start" (9), l'utilisateur peut démarrer des centrifugations mémorisées ou exécuter une centrifugation avec des paramètres présélectionnés manuellement. Lorsque les durées de fonctionnement présélectionnées sont terminées, la centrifugeuse s'arrêtera automatiquement. Avec la touche "Quick Spin" (8), l'utilisateur peut démarrer des centrifugations qui ne dureront que quelques secondes. Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-13 En appuyant sur la touche "Quick Spin" (8), la centrifugeuse accélère jusqu'à la rotation présélectionnée. Sur l'affichage "Time/Prog" (A-3), la durée de fonctionnement transmise est indiquée aussitôt que la touche "Quick Spin" (8) est enfoncée. En relâchant la touche "Quick Spin" (8), la centrifugeuse s'arrête et la durée de fonctionnement est indiquée jusqu'à ce que le couvercle soit ouvert. Figure. 26 3.6.2 La touche « STOP » Avec la touche "Stop" (10) (voir la figure 27), il est possible d'interrompre une centrifugation à tout moment. Après avoir appuyé sur la touche, la centrifugeuse décélère conformément à l'intensité présélectionnée jusqu'à son arrêt complet. Figure. 27 3.7 Détection d'un déséquilibrage Si la charge n'est pas également répartie sur le rotor, l'entraînement s'interrompra pendant l'accélération. Le rotor décélère jusqu'à l'arrêt complet. Lorsque sur l'affichage « Time/Prog" (A-3), le mot "error” (M11) apparaît conjointement au numéro "01”, la différence de poids des échantillons est trop importante. Répartir les poids uniformément. Charger le rotor selon la description des chapitres 3.1.2 et 3.1.3. Lorsque sur l'affichage « Time/Prog" (A-3), le mot "error” apparaît conjointement au numéro "02" (voir la figure 28), ceci peut être dû à la raison suivante : Le contacteur de déséquilibrage est défectueux. Figure. 28 FR-14 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4. PARAMÉTRAGE 4.1 Réglages de base 4.1.1 Accès au mode « Données de fonctionnement » Lors de l'utilisation de la centrifugeuse, les paramètres suivants doivent être définis : - Indication de la température en °C ou °F - Arrêt/Marche du signal acoustique - Arrêt/Marche du son du clavier - Présélection du volume sonore du signal - Sélection du signal sonore modulé "end of run" (centrifugation terminée) La centrifugeuse étant hors tension, appuyer simultanément sur les touches "Time"(6) et "Door Open" (7), et mettre la centrifugeuse sous tension avec le commutateur principal. Relâcher les deux touches ; il en résultera l'exécution d'un test de l'affichage pendant environ 5 secondes. Tous les indicateurs apparaîtront simultanément (voir la figure 29). Figure. 29 ATTENTION : L'utilisateur devra introduire le programme conformément à la description du paragraphe 4.1. 1 afin de modifier les réglages des paragraphes 4.1.2 - 4.1.7. Après avoir mémorisé les réglages, il est possible de revenir au mode de programmation normale en mettant hors tension la centrifugeuse pendant un court moment. Tous les réglages modifiés doivent être confirmés avec la touche "Start"(9). Le mot "Store"(12) apparaît sur l'affichage "RPM | RCF"(A-1) - Seules les présélections sont valides ! Figure. 30 4.1.2 Indication de la température Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" apparaît. Sélectionner maintenant la lettre "C" avec le bouton de réglage (1). Ceci entraîne l'apparition des mots "CELSI/temp" sur l'affichage "RPM | RCF" (A-1). En appuyant sur la touche "RPM | RCF" (4), le mot "CELSI" Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-15 clignote et il devient possible de modifier l'affichage en degrés Fahrenheit "FAREN" avec le bouton de réglage (1) (voir la figure 31). Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en mettant hors tension la centrifugeuse pendant un court moment. Figure. 31 4.1.3 Signal Marche/Arrêt Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" clignote. Sélectionner maintenant la lettre "L" avec le bouton de réglage (1). Ceci entraîne l'apparition des mots "On Sound" sur l'affichage "RPM | RCF" (4). En appuyant maintenant sur la touche "RPM | RCF" (4), le mot "On" clignote et il devient possible de couper le son avec le bouton de réglage (1) (voir la figure 32). Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en mettant hors tension la centrifugeuse pendant un court moment. Figure. 32 4.1.4 Volume de présélection du signal sonore Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" clignote. Sélectionner maintenant la lettre "U" avec le bouton de réglage (1). Ceci entraîne l'apparition des mots "Vol=09/Sound" sur l'affichage "RPM | RCF" (A-1). Après avoir appuyé sur la touche "RPM | RCF" (4), l'utilisateur peut régler le volume désiré entre 0 (faible) et 9 (fort) avec le bouton de réglage (1) (voir la figure 33). Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en mettant hors tension la centrifugeuse pendant un court moment. Figure. 33 FR-16 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4.1.5 Sélection de la sonorité du signal sonore de fin de centrifugation Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" clignote. Sélectionner maintenant la lettre "G" avec le bouton de réglage (1). Ceci entraîne l'apparition des mots "SonGo/Sound" sur l'affichage "RPM | RCF" (A-1). Après avoir appuyé sur la touche "RPM | RCF" (4), il est possible de sélectionner une sonorité avec le bouton de réglage (1). (voir la figure 34) Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en mettant hors tension la centrifugeuse pendant un court moment. Figure. 34 4.1.6 Marche/Arrêt du son du clavier Poursuivre conformément à la description du paragraphe 4.1.1 pour entrer dans ce mode de programme et appuyer ensuite sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" clignote. Sélectionner maintenant la lettre "B" avec le bouton de réglage (1). Ceci entraîne l'apparition des mots "ON/BEEP " sur l'affichage "RPM | RCF" (A-1). Après avoir appuyé sur la touche "RPM | RCF" (4), l'utilisateur peut mettre le son du clavier sur Marche ou sur Arrêt avec le bouton de réglage (1). (voir la figure 38) Après avoir mémorisé les réglages (voir 4.1.1), vous pouvez revenir au mode de programmation normale en mettant hors tension la centrifugeuse pendant un court moment. Figure. 35 4.1.7 Appel des données de fonctionnement ATTENTION : Ceci ne doit être réalisé que par des utilisateurs avertis ou par des techniciens de maintenance. Dans le mode « Réglages de base », il est possible de rappeler les données de fonctionnement de la centrifugeuse. Veuillez poursuivre conformément à la description du paragraphe 4.1. 1 pour entrer dans ce mode de programme. Appuyer sur la touche "Accel/Decel" (5). Sur l'affichage "Acc/Dec" (A-2), le mot "Service" clignote. Grâce au bouton de réglage (1), il est possible d'accéder à différentes informations : A = démarrages précédents de la centrifugeuse H = nombre d'heures de fonctionnement S = version du logiciel r = logiciel de conversion E = liste des messages d'erreur h = durée de fonctionnement du moteur Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-17 La liste des 99 derniers messages d'erreur peut être consultée en appuyant sur la touche "RPM | RCF" (4) et en faisant défiler la liste avec le bouton de réglage (1). Les codes d'erreurs apparaissent sur l'affichage "RPM | RCF" (A-1). Se reporter au "Tableau 6 : messages d'erreur (voir ANNEXE). Pour revenir au mode de programmation normale, mettre la centrifugeuse hors tension pendant un court moment. Figure. 36 5. MAINTENANCE 5.1 Maintenance et nettoyage 5.1.1 Entretien général : L’entretien de la centrifugeuse est limité au nettoyage du rotor, de la cuve du rotor et de ses accessoires, et à la lubrification régulière des boulons d'insertion de rotor pour un rotor horizontal (le cas échéant). Le lubrifiant préconisé est «High TEF Oil» Les lubrifiants contenant du disulfure de molybdène et du graphite ne sont pas autorisés. Prêter particulièrement attention aux pièces en aluminium anodisé. La rupture de rotors peut être provoquée par de très légères détériorations. Si des rotors, des godets ou des supports de tubes entrent en contact avec des substances corrosives, les surfaces concernées doivent être nettoyées avec précaution. Voici par exemple quelques substances corrosives : alcali, solution savonneuse alcaline, amines alcalines, acides concentrés, solution contenant des métaux lourds, solvants chlorés sans eau, solutions salines comme de l'eau salée, du phénol et des hydrocarbures halogénés. Nettoyage des unités, des rotors et des accessoires : - Mettre l'appareil hors tension et le déconnecter de l'alimentation électrique avant de commencer un nettoyage ou une désinfection. Ne pas verser de liquides à l'intérieur de l'enceinte. - Ne pulvériser du désinfectant sur l'appareil. - Un nettoyage complet est non seulement nécessaire à titre d'hygiène, mais aussi pour éviter une corrosion due à la pollution. - Afin d'éviter d'endommager des pièces anodisées comme les rotors, les plaques de réduction, etc., seuls des détergents à pH neutre (pH 6-8) peuvent être utilisés pour le nettoyage. Des produits de nettoyage alcalins (pH > 8) ne doivent pas être utilisés. - Après le nettoyage, s'assurer que toutes les pièces sont bien séchées soit manuellement, soit dans une armoire à air chaud (température maximum + 50 °C). - Il est nécessaire de revêtir régulièrement les pièces en aluminium anodisé avec de l'huile anti-corrosion afin d'augmenter leur durée de vie et de réduire les possibilités de corrosion. - De la condensation peut se for mer en présence d'humidité ou si les échantillons ne sont pas hermétiquement scellés. La condensation doit être régulièrement éliminée de la cuve du rotor avec un chiffon doux. La procédure de maintenance doit être répétée toutes les 10 à 15 centrifugations, ou au moins une fois par semaine. - Connecter l'unité à son alimentation après avoir complètement séché l'équipement. FR-18 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R - Aucune désinfection ne doit être effectuée avec des rayons ultraviolets, bêta et gamma, ou avec toute autre énergie rayonnante élevée. - Les rotors métalliques peuvent être passés à l'autoclave. - Le couvercle du rotor et les adaptateurs peuvent aussi être passés à l'autoclave (maxi. U21 °C, 20 min). - Les supports de tubes sont fabriqués en PP et ne peuvent pas être passés à l'autoclave à 134 °C. 5.1.2 Nettoyage et désinfection de l'unité 1. Ouvrir le couvercle avant de mettre l'unité hors tension. La déconnecter de son alimentation. 2. Dévisser l'écrou du rotor en tournant la clé du rotor dans le sens anti-horaire. 3. Démonter le rotor. 4. Nettoyer et désinfecter l'unité et la cuve du rotor en utilisant le nettoyant mentionné ci-dessus. 5. Nettoyer toutes les surfaces accessibles du dispositif et de ses accessoires, notamment le cordon électrique, avec un chiffon humide. 6. Laver abondamment les joints en caoutchouc et la cuve du rotor avec de l'eau. 7. Frotter les joints en caoutchouc sec avec du talc ou de la glycérine afin d'éviter qu'ils ne deviennent cassants. D'autres composants de l'appareil, comme le verrouillage du couvercle, l'arbre du moteur et le rotor, ne doivent pas être graissés. 8. Sécher l'arbre du moteur avec un chiffon doux, sec et non pelucheux. 9. Vérifier l'unité et les accessoires à la recherche de détériorations. S'assurer que la centrifugeuse est hors tension et déconnecter l'unité de l'alimentation électrique. Éliminer ensuite la poussière adhérant aux fentes d'aération dans la centrifugeuse avec une brosse souple. Ceci doit être réalisé au moins tous les six mois. 5.1.3 Nettoyage et désinfection du rotor 1. Nettoyer et désinfecter les rotors, les couvercles des rotors ainsi que les adaptateurs avec le nettoyant mentionné ci-dessus. 2. Utiliser un goupillon pour nettoyer et désinfecter les alésages du rotor. 3. Rincer les rotors, le couvercle du rotor et l'adaptateur à l'eau claire. Particulièrement les passages perforés des rotors angulaires. 4. Pour sécher les rotors et les accessoires, les installer sur une serviette. Positionner les rotors angulaires avec les alésages vers le bas. 5. Sécher le cône du rotor avec un chiffon doux, sec et non pelucheux, tout en recherchant d'éventuelles détériorations. Ne pas graisser le cône du rotor. 5.1.4 Désinfection des rotors en aluminium Si des produits infectieux se répandent dans la centrifugeuse, le rotor et la cuve du rotor doivent être désinfectés immédiatement après la centrifugation. Les rotors peuvent être passés à l'autoclave à une température maximum de 121 °C. 5.1.5 Désinfection des rotors PP Autoclave Durée recommandée de passage à l'autoclave : 15 – 20 min à 121 °C (1 bar) ATTENTION : La durée de stérilisation de 20 minutes ne doit pas être dépassée. Des stérilisations répétées provoqueront la diminution de la résistance mécanique des matériaux en plastique Avant de passer le rotor PP et l'adaptateur à l'autoclave, ils doivent être bien nettoyés afin d'éviter de « cuire » les résidus de saletés. Ne pas tenir compte des effets de certains résidus chimiques sur les matériaux en plastique aux températures ambiantes. Mais avec les températures élevées en autoclave, ces résidus peuvent corroder et détruire le plastique. Les objets doivent être abondamment rincés à l'eau distillée après le nettoyage mais avant de les passer à l'autoclave. Les résidus de liquide de nettoyage peuvent provoquer des fissures, un blanchissement et des taches. Stérilisation au gaz Adaptateurs, bouteilles et rotors sont stérilisables au gaz avec de l'oxyde d'éthylène. S'assurer de laisser les articles à l'air libre après la stérilisation et avant de les réutiliser. ATTENTION : La température pouvant augmenter pendant la stérilisation, les rotors, les adaptateurs et les bouteilles ne doivent pas être fermés et doivent être complètement dévissés. Stérilisation chimique Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-19 Bouteilles, adaptateurs et rotors peuvent être traités avec des désinfectants liquides usuels. ATTENTION : Avant d'appliquer toute autre méthode de nettoyage ou de décontamination que celle recommandée par le fabricant, contacter ce dernier afin d'être assuré qu'elle ne détériorera pas l'unité ou le rotor. 5.1.6 Bris de verre Avec des valeurs d'accélération élevée, le taux de bris de tubes en verre augmente. Des éclats de verre doivent être immédiatement éliminés du rotor, des godets, des adaptateurs et de la cuve du rotor elle-même. De fins éclats de verre rayeront et endommageront le revêtement protecteur de surface d'un rotor. Si des éclats de verre restent dans la cuve du rotor, une fine poussière métallique s'accumulera en raison de la circulation d'air. Cette très fine poussière métallique noire polluera fortement la cuve du rotor, le rotor, les godets et les échantillons. Remplacer au besoin les adaptateurs, les tubes et les accessoires afin d'éviter d'amplifier les détériorations. Vérifier régulièrement les alésages du rotor à la recherche de résidus et de détériorations. ATTENTION : Vérifier les spécifications particulières des tubes avec le fabricant. Durée d'utilisation des rotors, des godets et des accessoires Les rotors et le couvercle de rotor en aluminium ou en acier inoxydable ont une durée opérationnelle de 7 ans au maximum après la première utilisation. Les bouchons et les couvercles de rotor transparents fabriqués en PC ou en PP ainsi que les rotors, les supports de tubes et les adaptateurs en PP ont une durée de fonctionnement maximum de 3 ans après la première utilisation. Conditions de la durée de fonctionnement : Utilisation adéquate, protections en place et entretien recommandé. 6. DÉPANNAGE 6.1 Message d'erreur : Causes/solutions Les messages d'erreur sont répertoriés afin d'aider à localiser plus rapidement de possibles erreurs. Le diagnostic de référence dans ce chapitre peut ne pas toujours correspondre à la situation dans la mesure où les erreurs et les solutions proposées sont théoriques. Veuillez nous informer de toutes les sortes d'erreurs se produisant qui ne sont pas répertoriées dans ce chapitre. C'est grâce à vos informations en retour que nous sommes en mesure d'améliorer ce manuel de fonctionnement. Nous vous remercions dès maintenant de votre aide. 6.2 Étude des messages d'erreur possibles et leurs solutions 6.2.1 Libération du couvercle pendant une panne d'alimentation (libération en cas d'urgence) En cas de panne d'alimentation ou de défaut de fonctionnement, le couvercle de la centrifugeuse peut être ouvert manuellement afin récupérer les échantillons. Pour les modèles FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (moteur de verrouillage entraîné) Poursuivre de la manière suivante : 1. Mettre la centrifugeuse hors tension, débrancher le cordon d'alimentation et attendre que le rotor soit à l'arrêt complet (ce qui peut prendre plusieurs minutes) 2. Sur le côté gauche de l'enceinte de la centrifugeuse se trouve une butée en plastique. Retirer la butée derrière laquelle se trouve un écrou hexagonal. 3. Utiliser la clé polygonale incluse, l'introduire dans l'orifice et serrer la clé polygonale avec l'écrou hexagonal (voir la figure 37). 4. Faire ensuite tourner la clé polygonale vers la droite (dans le sens horaire) jusqu'à la limite. ATTENTION : a) Ne tourner que jusqu'à la limite sans serre r l'écrou. b) Il est dorénavant possible d'ouvrir le couvercle de la centrifugeuse. c) Remettre la centrifugeuse sous tension pour reprendre le travail. FR-20 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figure. 37 Pour 5714 Procéder comme suit (cf. Figure 38): ATTENTION: • Arrêter la centrifugeuse et débrancher le cordon électrique, attendre que le rotor soit à l’arrêt (cela peut prendre plusieurs minutes). Sur le côté droit, il y a un bouchon en plastique (Figure 38). Ôter le bouchon, raccordé au verrou du capot, horizontalement du corps, jusqu’à ce que le capot de la centrifugeuse soit ouvert. • Ouvrir maintenant le capot de la centrifugeuse. Figure 38 6.2.2 Description du système de messages d'erreur Le message d'erreur "error" (M11) apparaît sur l'affichage "Time/Prog" (A-3) (voir la figure 39). Des informations détaillées sur les messages d'erreur possibles se trouvent sur le « tableau 6 : messages d'erreur" (voir Annexe). Figure. 39 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-21 7. RÉCEPTION DES CENTRIFUGEUSES POUR RÉPARATION Danger pour la santé par des équipements, des rotors et des accessoires contaminés. En cas de retour d'une centrifugeuse pour réparation auprès du fabricant, prendre note de ce qui suit : La centrifugeuse doit être décontaminée et nettoyée avant expédition afin de protéger les personnes, l'environnement et le matériel. Certificat de décontamination pour le retour de matériel (voir ANNEXE) Le fabricant se réserve le droit de ne pas accepter de centrifugeuses contaminées. De plus, toutes les dépenses de nettoyage et de désinfection des unités seront débitées du compte du client 8. TRANSPORT, STOCKAGE ET MISE AU REBUT 8.1 Transport Avant le transport, extraire le rotor. Ne transporter cette unité que dans son emballage d'origine. Utiliser une aide au transport pour une expédition sur de longues distances afin de bloquer l'arbre du moteur. Généralités sur le transport Température de l'air -25 à 60 ° Humidité relative 10 à 75 % Pression d'air 30 à 106 kPa 8.2 Stockage Pendant le stockage de la centrifugeuse, les conditions environnementales suivantes doivent être observées : dans l'emballage de transport Température de l'air -25 à 55 ° Humidité relative 10 à 75 % Pression d'air 70 à 106 kPa 8.3 Transport, installation, transfert et mise au rebut de la centrifugeuse FC5714/FC5718/FC5718R/FC5816/FC5816R Ces instructions viennent en complément des instructions générales précédentes du chapitre 8 et ne les remplacent pas. 8.3.1 Transport • Transporter l'équipement dans son emballage d'origine. • La centrifugeuse doit toujours être transportée par deux personnes. • Utiliser une aide au transport pour une expédition sur de plus longues distances. 8.3.2 Installation La centrifugeuse doit toujours être transportée par deux personnes.  Ouverture du carton et extraction de l'équipement. 1. Couper la bande adhésive. 2. Ouvrir les 4 rabats du carton. 3. Enlever les accessoires. 4. Placer les mains sous l'équipement et soulever la centrifugeuse hors du carton avec une autre personne. FR-22 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R  Placer l'équipement sur une table de laboratoire stable, horizontale et anti-vibratoire 1. Enlever le matériau de protection à l'avant et à l'arrière. 2. Enlever le manchon en plastique. 3. Respecter une distance minimum de 30 cm sur les côtés (dispositifs adjacents) et à l'arrière (mur). 4. Installer l'équipement dans un endroit bien aéré, protégé de la lumière directe du soleil pour éviter toute surchauffe.  Connexion de l'équipement 1. Après l'installation, attendre quatre heures avant de brancher la centrifugeuse afin d'éviter d'endommager le compresseur. 2. Vérifier que la tension et la fréquence du réseau électrique correspondent aux spécifications sur la plaque de l'équipement (voir l'arrière du dispositif), puis connecter l'équipement à l'alimentation électrique.  Enlever le matériau de protection de la cuve du rotor 1. Brancher l'équipement sur l'interrupteur d'alimentation. 2. Ouvrir le couvercle de la centrifugeuse à l'aide du bouton d'ouverture. 4. Enlever le matériau de protection utilisé pour le transport. 5. Placer le rotor à la verticale sur l'arbre du moteur. 6. Tourner l'écrou du rotor dans le sens horaire à l'aide de la clé de rotor jusqu'à ce que l'écrou soit bien serré.  L'équipement est maintenant prêt à l'utilisation Conserver l'emballage et tous les matériaux de protection en cas d'expédition ultérieure de l'équipement. 8.3.3 Emballage Emballer la centrifugeuse dans l'ordre inverse. 8.3.4 Transfert de l'équipement Lors du passage de l'équipement à des tiers, s'il vous plaît assurez-vous d'inclure également ce mode d'emploi. Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9. DONNEES TECHNIQUES 9.1 Spécifications 9.1.1 Centrifugeuse FC5714 Modèle Plage des v itesses Maximum RCF Capacité maximum (rotor) Plage de température Durée de fonctionnement Niv eau du bruit (en fonction du rotor) Densité acceptable à la v itesse maximum Énergie cinétique acceptable Connexion électrique CA Variations de tension Consommation électrique Consommation Dimensions (L × P × H) Poids net (sans rotor) Dimensions d'expédition (L × P × H) Poids à l'expédition (sans rotor) FC5714 200 rpm -14000 rpm;10 rpm/set 18624 x g;10 x g/set 4x100ml Air froid 10 sec à 99 h 59 min 59 sec ou en continu ≤ 63 ± 2 dB(A) 1.2 g/ml 5595 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % 1.3 A 2.4 A 240 W 300 W 362 x 493 x 330 mm 14.3 x 19.4 x 13.0 in 30 kg 66 lb 580 x 490 x 460 mm 22.8 x 19.3 x 18.1 in 32.5 kg 72 lb Conditions ambiantes (EN/IEC 61010-1) Env ironnement Altitude Température ambiante Humidité relativ e maximum Catégorie de surtension (IEC 60364-4-443) Niv eau de contamination Classe de protection pour utilisation à l’intérieur uniquement Utilisation jusqu'à une altitude maximum de 2000 mètres 2°C jusqu'à 35 °C Humidité relative maximum de 80 % pour des températures jusqu'à 31 °C, avec diminution linéaire jusqu'à 50 % de l'humidité relative à 35 °C. Inadaptée à une utilisation dans des endroits dangereux. II 2 I FR-23 FR-24 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.2 Centrifugeuse FC5718 Modèle Plage des v itesses Maximum RCF Capacité maximum (rotor) Plage de température Durée de fonctionnement Niv eau du bruit (en fonction du rotor) Densité acceptable à la v itesse maximum Énergie cinétique acceptable Connexion électrique CA Variations de tension Consommation électrique Consommation FC5718 200 rpm -18000 rpm;10 rpm/set 23542 x g;10 x g/set 4x100ml Air froid 10 sec à 99 h 59 min 59 sec ou en continu ≤ 60 ± 2 dB(A) 1.2 g/ml 16672 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % 2.0 A 4.0 A 455 W 475 W 408 x 499 x 351 mm 16.1 x 19.7 x 13.8 in 43 kg 95 lb 650 x 520 x 490 mm 25.6 x 20.5 x 19.3 in 53 kg 117 lb Dimensions (L × P × H) Poids net (sans rotor) Dimensions d'expédition (L × P × H) Poids à l'expédition (sans rotor) Conditions ambiantes (EN/IEC 61010-1) Env ironnement Altitude Température ambiante Humidité relativ e maximum pour utilisation à l’intérieur uniquement Utilisation jusqu'à une altitude maximum de 2000 mètres 2 °C jusqu'à 35 °C Humidité relative maximum de 80 % pour des températures jusqu'à 31 °C, avec diminution linéaire jusqu'à 50 % de l'humidité relative à 35 °C. Catégorie surtension (IEC 60364-4-443) Niv eau de contamination Classe de protection Inadaptée à une utilisation dans des endroits dangereux. II 2 I Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.3 Centrifugeuse FC5718R Modèle Plage des v itesses Maximum RCF Capacité maximum (rotor) Plage de température Durée de fonctionnement Niv eau du bruit (en fonction du rotor) Densité acceptable à la v itesse maximum Énergie cinétique acceptable Connexion électrique CA Variations de tension Consommation électrique Consommation FC5718R 200 rpm -18000 rpm;10 rpm/set 23542 x g;10 x g/set 4x100ml -20~40℃ 10 sec à 99 h 59 min 59 sec ou en continu ≤ 60 ± 2 dB(A) 1.2 g/ml 25111 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % 3.0 A 6.0 A 660 W 660 W 407 x 731 x 359 mm 16.0 x 28.8 x 14.1 in 60 kg 132 lb 840 x 640 x 590 mm 33.1 x 25.2 x 23.2 in 77 kg 170 lb Dimensions (L × P × H) Poids net (sans rotor) Dimensions d'expédition (L × P × H) Poids à l'expédition (sans rotor) Conditions ambiantes (EN/IEC 61010-1) Env ironnement Altitude Température ambiante Humidité relativ e maximum pour utilisation à l’intérieur uniquement Utilisation jusqu'à une altitude maximum de 2000 mètres 2 °C jusqu'à 35 °C Humidité relative maximum de 80 % pour des températures jusqu'à 31 °C, avec diminution linéaire jusqu'à 50 % de l'humidité relative à 35 °C. Catégorie surtension (IEC 60364-4-443) Niv eau de contamination Classe de protection Inadaptée à une utilisation dans des endroits dangereux. II 2 I FR-25 FR-26 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.4 Centrifugeuse FC5816 Modèle Plage des v itesses Maximum RCF Capacité maximum (rotor) Plage de température Durée de fonctionnement Niv eau du bruit (en fonction du rotor) Densité acceptable à la v itesse maximum Énergie cinétique acceptable Connexion électrique CA Variations de tension Consommation électrique Consommation FC5816 200 rpm -15000 rpm;10 rpm/set 21379 x g;10 x g/set 6 x 250ml Air froid 10 sec à 99 h 59 min 59 sec ou en continu ≤ 61 ± 2 dB(A) 1.2 g/ml 34363 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % 2.4 A 4.2 A 530 W 520 W 446 x 538 x 354 mm 17.6 x 21.2 x 13.9 in 52 kg 115 lb 840 x 640 x 590 mm 33.1 x 25.2 x 23.2 in 77 kg 170 lb Dimensions (L × P × H) Poids net (sans rotor) Dimensions d'expédition (L × P × H) Poids à l'expédition (sans rotor) Conditions ambiantes (EN/IEC 61010-1) Env ironnement Altitude Température ambiante Humidité relativ e maximum pour utilisation à l’intérieur uniquement Utilisation jusqu'à une altitude maximum de 2000 mètres 2 °C jusqu'à 35 °C Humidité relative maximum de 80 % pour des températures jusqu'à 31 °C, avec diminution linéaire jusqu'à 50 % de l'humidité relative à 35 °C. Catégorie surtension (IEC 60364-4-443) Niv eau de contamination Classe de protection Inadaptée à une utilisation dans des endroits dangereux. II 2 I Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.5 Centrifugeuse FC5816R Modèle Plage des v itesses Maximum RCF Capacité maximum (rotor) Plage de température Durée de fonctionnement Niv eau du bruit (en fonction du rotor) Densité acceptable à la v itesse maximum Énergie cinétique acceptable Connexion électrique CA Variations de tension Consommation électrique Consommation FC5816R 200 rpm -16000 rpm;10 rpm/set 24325 x g;10 x g/set 6 x 250ml -20~40℃ 10 sec à 99 h 59 min 59 sec ou en continu ≤ 63 ± 2 dB(A) 1.2 g/ml 34363 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % 3.7 A 7.8 A 785 W 850 W 723 x 538 x 354 mm 28.5 x 21.2 x 13.9 in 77 kg 170 lb 840 x 640 x 590 mm 33.1 x 25.2 x 23.2 in 87 kg 192 lb Dimensions (L × P × H) Poids net (sans rotor) Dimensions d'expédition (L × P × H) Poids à l'expédition (sans rotor) Conditions ambiantes (EN/IEC 61010-1) Env ironnement Altitude Température ambiante Humidité relativ e maximum pour utilisation à l’intérieur uniquement Utilisation jusqu'à une altitude maximum de 2000 mètres 2 °C jusqu'à 35 °C Humidité relative maximum de 80 % pour des températures jusqu'à 31 °C, avec diminution linéaire jusqu'à 50 % de l'humidité relative à 35 °C. Catégorie surtension (IEC 60364-4-443) Niv eau de contamination Classe de protection Inadaptée à une utilisation dans des endroits dangereux. II 2 I FR-27 FR-28 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.6 Centrifuge FC5916 Modèle FC5916 200 tr/min -16000 tr/min;10 tr/min/set Plage de v itesse 24325 x g;10 x g/set ACR max. 4x750ml Capacité maxi(Rotor) Plage de température(N/A) Refroidissement par air Durée de fonctionnement 10 sec à 99 h 59 min 59 sec ou en continu ≤ 63 ± 2 dB(A) Niv eau sonore (selon le rotor) Densité admissible à la v itesse maximale 1.2 g/ml 60629 Nm Énergie cinétique admissible Raccordement au réseau AC Fluctuation de la tension 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % Consommation de courant 2.8 A 5.6 A Consommation d'énergie 640 W 680 W 540 x 670 x 390 mm Dimensions (L × P × H) 21.3 x 26.4 x 15.4 in 85 kg Poids net (sans rotor) 187 lb 780 x 670 x 590 mm Dimensions d'expédition (L × P × H) 30.7 x 26.4 x 23.3 in 98 kg Poids à l'expédition (sans rotor) 216 lb Conditions ambiantes (EN/IEC 61010-1) Env ironnement Altitude Température ambiante Pour usage intérieur seulement Utilisation jusqu'à une altitude de 2000 m 2°C jusqu'à 35 °C Humidité relative max. 80 % pour des températures jusqu'à 31°C, Humidité relativ e max. diminuant linéairement jusqu'à 50 % d'humidité relative jusqu'à 35°C. Catégorie de surtension (IEC 60364-4-443) Degré de contamination Classe de protection Ne conv ient pas à une utilisation dans des env ironnements dangereux. II 2 I Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.7 Centrifuge FC5916R Modèle FC5916R 200 rpm -16000 rpm;10 rpm/set Plage de v itesse 26331 x g;10 x g/set ACR max. Capacité maxi(Rotor) 4x750ml Plage de température(Digitale) -20~40℃ 10 sec à 99 h 59 min 59 sec ou en continu Durée de fonctionnement ≤ 63 ± 2 dB(A) Niv eau sonore (selon le rotor) Densité admissible à la v itesse maximale 1.2 g/ml 54458 Nm Énergie cinétique admissible Raccordement au réseau AC Fluctuation de la tension Consommation de courant Consommation d'énergie 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % 7.2 A 20 A 1630 W 1750 W 730 x 670 x 390 mm Dimensions (L × P × H) 28.7 x 26.4 x 15.4 in 118 kg Poids net (sans rotor) 260 lb 900 x 750 x 560 mm Dimensions d'expédition (L × P × H) 40.0 x 29.5 x 22.0 in 137 kg Poids à l'expédition (sans rotor) 137 kg Conditions ambiantes (EN/IEC 61010-1) Env ironnement Altitude Température ambiante Pour usage intérieur seulement Utilisation jusqu'à une altitude de 2000 m 2°C jusqu'à 35 °C Humidité relative max. 80 % pour des températures jusqu'à 31°C, Humidité relativ e max. diminuant linéairement jusqu'à 50 % d'humidité relative jusqu'à 35°C. Catégorie de surtension (IEC 60364-4-443) II Degré de contamination 2 Classe de protection I Ne conv ient pas à une utilisation dans des env ironnements dangereux. FR-29 FR-30 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.2 Schémas et dimensions D H W D H W Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Modèle FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R L (mm / in.) 362 / 14.3 408 / 16.1 408 / 16.1 446 / 17.6 723 / 28.5 540 / 21.3 730 / 28.7 P (mm / in.) 493 / 19.4 499 / 19.7 731 / 28.8 538 / 21.2 538 / 21.2 670 / 26.4 670 / 26.4 H (mm / in.) 330 / 13.0 351 / 13.8 359 / 14.1 354 / 13.9 354 / 13.9 390 / 15.4 390 / 15.4 10. CONFORMITÉ La conformité aux normes suivantes est indiquée par le marquage correspondant sur le produit. Marquage Standard Ce produit est conforme aux normes harmonisées applicables des directives européennes 2011/65/UE (RoHS), 2014/30/UE (CEM) et 2014/35/UE (LVD). La déclaration de conformité de l'UE est disponible en ligne à www.ohaus.com/ce. Mise au rebut Conformément à la directive européenne 2012/19/UE sur les déchets d’équipements électriques et électroniques (WEEE), cet appareil ne doit pas être mis au rebut avec les ordures ménagères. Ceci est également valable pour les pays en dehors de l’UE, selon leurs spécifications particulières. Veuillez mettre au rebut ce produit conformément à la réglementation locale au point de collecte spécifié pour les équipements électriques et électroniques. En cas de questions, veuillez contacter l’autorité responsable ou le distributeur auprès duquel cet appareil a été acheté. Si cet appareil change de propriétaire (pour des raisons personnelles ou professionnelles), cette consigne doit lui être communiquée. Pour les instructions de rejet et recyclage en Europe, merci de consulter le site www.ohaus.com/weee. Nous vous remercions de votre contribution à la protection de l’environnement. FR-31 FR-32 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11. ANNEXE TABLEAU 1 : POIDS NET AUTORISE TABLEAU 2 : TEMPERATURES LES PLUS FAIBLES A LA VITESSE MAXIMUM TABLEAU 3 : VITESSE ET VALEURS RCF MAXI POUR LES ROTORS AUTORISES TABLEAU 4 : DUREES D'ACCELERATION ET DE DECELERATION TABLEAU 5 : MESSAGES D'ERREUR TABLEAU 6: CORRECTION DU RAYON TABLEAU 7 : FORMULAIRE DE RACHAT/CERTIFICAT DE DECONTAMINATION Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-33 11.1 Table 1: POIDS NET AUTORISE FC5916R FC5916 FC5816R FC5816 Poids net autorisé FC5718R Description du rotor FC5718 Numé ro de rotor FC5714 Vitesse maxi ( x1000 tr/min) 14.0 14.0 15.0 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 24x 3.4 g 15.0 16.0 16.0 16.0 30314820 Rotor, Sw ing out, 4x 250ml, ID 4.5 4x 536 g 4.5 8.0 30314821 Rotor, Angle, 6x 250ml FB, ID 8.0 6x 355 g 8.0 8.0 4.5 30314822 Rotor, Sw ing out, 4x 100ml, ID 4x 340 g 4.5 4.5 4.0 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 4x 465 g 5.0 5.0 4.5 4.5 4.5 30314824 Rotor, Sw ing out, 2x 3MTP, ID 2x 310 g 4.5 4.5 4.5 4.5 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 6x 140 g 11.0 13.5 9.0 9.0 30314826 Rotor, Angle, 6x 85ml RB, ID 11.0 6x 140 g 13.0 11.0 11.0 12.0 12.0 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 4x 140 g 12.0 12.0 15.0 15.0 30314828 Rotor, Sw ing out, 16x 50ml, ID 4x 557 g 4.5 4.5 FR-34 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 7.5 7.5 30314829 Rotor, Angle, 10x 50ml FA, ID 9.0 10x 74 g 10.5 10.0 10.0 6.0 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 6x 72 g 6.0 6.0 12.0 12.0 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 13.0 6x 94 g 13.0 13.0 13.0 4.5 4.5 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30x 32 g 4.5 4.5 4.5 12.0 12.0 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 12.0 20x 18 g 12.0 12.0 12.0 6.0 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 12x 25 g 6.0 6.0 13.5 13.5 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 15.0 44x 3.8 g 16.0 16.0 16.0 12.0 13.0 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30x 3.4 g 14.0 15.0 15.0 14.0 14.0 15.0 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 24x 3.4 g 15.0 16.0 15.0 15.0 30314839 30314840 Rotor, Angle, 12x 1.5/2.0ml, ID Rotor, Angle, 64x 0.5ml, ID 12x 3.4 g 64x 2.3 g 18.0 18.0 13.5 13.5 15.0 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 4x 3.5 g 15.0 15.0 15.0 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID 4x 995 g 4x 708 g 4.0 4.0 8.0 8.0 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-35 11.2 Table 2: TEMPERATURES LES PLUS FAIBLES A LA VITESSE MAXIMUM Numé ro de rotor Description du rotor 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 30314840 Rotor, Angle, 64x 0.5ml, ID 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID Modè le FC5718R FC5816R FC5916R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5916R FC5916R Vitesse maximale 15000 rpm 15000 rpm 16000 rpm 8000 rpm 4500 rpm 4500 rpm 5000 rpm 5000 rpm 4500 rpm 4500 rpm 4500 rpm 13500 rpm 13500 rpm 9000 rpm 9000 rpm 13000 rpm 12000 rpm 12000 rpm 16000 rpm 7500 rpm 7500 rpm 10500 rpm 6000 rpm 6000 rpm 12000 rpm 12000 rpm 13000 rpm 4500 rpm 4500 rpm 12000 rpm 12000 rpm 12000 rpm 6000 rpm 6000 rpm 13500 rpm 13500 rpm 16000 rpm 14000 rpm 14000 rpm 15000 rpm 15000 rpm 15000 rpm 16000 rpm 18000 rpm 18000 rpm 13500 rpm 13500 rpm 15000 rpm 15000 rpm 15000 rpm 4500 rpm 8000 rpm N-max 6 °C 6 °C 1 °C -1 ° C -7 ° C -7 ° C -3 ° C -3 ° C -6 ° C -6 ° C -8 ° C 17 ° C 17 ° C -5 ° C -5 ° C 8 °C 5 °C 5 °C 7 °C 9 °C 9 °C 2 °C -8 ° C -8 ° C -5 ° C -5 ° C -4 ° C -9 ° C -9 ° C 2 °C 2 °C -2 ° C -11.5 ° C -11.5 ° C 1 °C 1 °C 3 °C 9 °C 9 °C 3 °C 6 °C 6 °C 1 °C -2 ° C -2 ° C 2 °C 2 °C -1 ° C -1 ° C -2 ° C 4 °C 0 °C Toutes les indications de température se réfèrent à une température ambiante de 23 ° C. En dépassant cette valeur ou le rayonnement solaire direct vers la centrifugeuse, ces valeurs ne peuvent pas être maintenues. FR-36 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.3 Table 3: VITESSE ET VALEURS RCF MAXI POUR LES ROTORS AUTORISES Numé ro de rotor 30304361 30314820 30314821 30314822 30314823 30314824 30314825 30314826 30314827 30314828 Description du rotor Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable Rotor, Sw ing out, 4x 250ml, ID Rotor, Angle, 6x 250ml FB, ID Rotor, Sw ing out, 4x 100ml, ID Rotor, Sw ing out, 4x 100ml, ID, Sealable Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID Rotor, Angle, 4x 85ml RB, ID, Hi Rotor, Sw ing out, 16x 50ml, ID Modè le Vitesse maximale RCF 18624 x g FC5714 RPM 14000 FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5816 4500 3780 x g FC5816R 4500 3780 x g FC5816 8000 10016 x g FC5816R 8000 10016 x g FC5916 8000 10016 x g FC5916R 8000 10016 x g FC5714 4500 3350 x g FC5718 4500 3350 x g FC5718R 4500 3350 x g FC5714 4000 2486 x g FC5718 5000 3885 x g FC5718R 5000 3885 x g FC5714 4500 2716 x g FC5718 4500 2716 x g FC5718R 4500 2716 x g FC5816 4500 2716 x g FC5816R 4500 2716 x g FC5916 4500 2716 x g FC5916R 4500 2716 x g FC5718 11000 13932 x g FC5718R 13500 20984 x g FC5718 9000 10413 x g FC5718R 9000 10413 x g FC5816 11000 15555 x g FC5816R 13000 21726 x g FC5916 11000 15555 x g FC5916R 13000 21726 x g FC5718 12000 14809 x g FC5718R 12000 14809 x g FC5816 12000 14809 x g FC5816R 12000 14809 x g FC5916 15000 23140 x g FC5916R 16000 26328 x g FC5816 4500 3735 x g FC5816R 4500 3735 x g Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314829 30314830 30314831 30314832 30314833 30314834 30314835 30314836 Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Rotor, Angle, 20x 10ml RB, ID, Hi Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable FR-37 FC5718 7500 8174 x g FC5718R 7500 8174 x g FC5816 9000 11771 x g FC5816R 10500 16022 x g FC5916 10000 14532 x g FC5916R 10500 16022 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 12000 13522 x g FC5718R 12000 13522 x g FC5816 13000 15869 x g FC5816R 13000 15869 x g FC5916 13000 15869 x g FC5916R 13000 15869 x g FC5714 4500 2830 x g FC5718 4500 2830 x g FC5718R 4500 2830 x g FC5816 4500 2830 x g FC5816R 4500 2830 x g FC5718 12000 15775 x g FC5718R 12000 15775 x g FC5816 12000 15775 x g FC5816R 12000 15775 x g FC5916 12000 15775 x g FC5916R 12000 15775 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 13500 17113 x g FC5718R 13500 17113 x g FC5816 15000 21128 x g FC5816R 16000 24039 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5714 12000 15131 x g FC5718 13000 17758 x g FC5718R 14000 20595 x g FC5916 15000 23643 x g FC5916R 15000 23643 x g FR-38 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 30314840 30314841 30553085 30553086 Rotor, Angle, 12x 1.5/2.0ml, ID Rotor, Angle, 64x 0.5ml, ID Rotor, Angle, 4x 8-w PCR Strip, ID Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID FC5714 14000 18624 x g FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5718 18000 23542 x g FC5718R 18000 23542 x g FC5718 13500 16298 x g FC5718R 13500 16298 x g FC5718 15000 15343 x g FC5718R 15000 15343 x g FC5916 15000 15343 x g FC5916R 15000 15343 x g FC5916 4000 3434 x g FC5916R 4500 4346 x g FC5916 8000 10367 x g FC5916R 8000 10367 x g Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R FR-39 11.4 Table 4: DUREES D'ACCELERATION ET DE DECELERATION N° d'article Description du rotor Modè le Accé lé ration* Temps en sec niveau 0 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314820 Rotor, Sw ing out, 4x 250ml, ID 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314828 Rotor, Sw ing out, 16x 50ml, ID 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R 200 240 230 251 249 204 204 309 309 664 664 573 573 110 100 100 110 100 100 230 160 160 452 452 249 249 400 500 360 360 697 2313 463 549 300 300 506 506 448 448 311 311 380 380 753 753 480 480 101 89 90 niveau 9 25 23 25 25 20 21 21 34 34 130 130 66 66 14 15 15 14 15 15 25 20 20 43 43 27 27 40 60 40 40 85 70 48 69 60 60 60 60 50 50 34 34 63 62 115 115 60 60 11 13 12 Dé cé lé ration* Temps en sec niveau 0 230 240 420 610 635 421 421 458 458 2906 2906 1903 1903 170 150 150 170 150 150 340 360 360 616 616 488 488 960 1260 1050 1050 2313 1630 1654 1307 820 820 1745 1745 1251 1251 387 387 796 1069 2395 2395 1747 1747 206 239 463 niveau 9 35 16 17 26 27 30 30 36 36 92 92 84 84 19 15 15 17 15 15 26 15 15 38 38 23 23 40 50 40 40 70 76 46 67 40 40 49 49 45 45 36 36 37 36 72 72 68 68 14 11 11 FR-40 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 30314840 Rotor, Angle, 64x 0.5ml, ID 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 30553085 Rotor Sw ingOut 4x 750ml D99mm ID SL 30553086 Rotor Angle 4x 500ml D69mm ID FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5916 FC5916R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5718 FC5718R FC5916 FC5916R FC5916 FC5916R FC5916 FC5916R *Note: accelerates from 0 to Vmax; decelerates from Vmax to 0. 360 360 446 446 264 264 157 160 160 149 149 360 360 512 512 305 305 102 88 88 196 196 257 274 233 233 250 200 210 221 221 200 210 230 251 249 204 204 240 240 150 150 130 130 100 100 483 551 575 575 40 40 48 48 28 28 19 15 15 25 25 40 40 54 54 32 32 11 13 12 22 22 29 31 25 25 27 25 25 23 23 22 23 25 25 30 21 21 25 25 17 17 16 16 12 12 47 55 73 73 570 570 1323 1323 921 921 370 380 380 985 985 570 570 1439 1439 988 988 167 222 433 204 204 436 485 283 500 280 360 360 561 561 230 240 420 610 635 421 421 210 210 170 170 130 130 201 201 1287 1501 2317 2317 30 30 49 49 32 32 20 10 10 20 20 30 30 47 47 37 37 14 11 11 18 18 28 29 30 25 34 25 30 30 30 35 16 17 26 27 30 30 20 20 12 12 12 12 12 12 49 54 82 82 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.5 Tableau 5 : MESSAGES D'ERREUR N° d'erreur : Description 1 2 4 8 11 12 14 FERMETURE du couvercle 33 34 38 40 41 42 43 44 45 46 47 48 49 55 70 99 FAUX rotor n° Survenue d'un dé sé quilibrage Le capteur de dé sé quilibrage est dé fectueux Le contacteur de dé sé quilibrage a é té activé pendant plus de 5 secondes Le transpondeur dans le rotor est dé fectueux Le capteur de tempé rature est dé fectueux Surchauffe de la cuve La variation de vitesse est trop importante entre deux mesures Ouverture du couvercle lorsque le moteur est en fonctionnement Contact du couvercle dé fectueux Le moteur du couvercle est bloqué Les communications avec le convertisseur de fré quence sont perturbé es pendant le dé marrage Les communications avec le convertisseur de fré quence sont perturbé es pendant l'arrê t Court-circuit dans le convertisseur de fré quence Tension trop faible sur le convertisseur de fré quence Surtension sur le convertisseur de fré quence Surchauffe du convertisseur de fré quence Surchauffe du moteur Surintensité du convertisseur de fré quence Temporisation entre l'unité de contrô le et le convertisseur de fré quence Autre erreur du convertisseur de fré quence Survitesse Temporisation entre le contrô leur et l'interface RS232 Rotor non autorisé dans cette centrifugeuse Le rotor insé ré n'existe pas dans le programme Le rotor n'est pas dé tecté FR-41 FR-42 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.6 Table 6: CORRECTION DU RAYON Compatible FC5916 FC5916R Blank NO Bucket/Adapter 8.5 ● ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.3 ● ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 -1.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 16.7 ● ● 30314903 Rack, 1x 250ml D62mm FB, 2/pk 15.7 -1.0 ● ● 30314904 Rack, 1x 100ml D41mm RB, 2/pk 15.9 -0.8 ● ● 30314905 Rack, 3x 50ml D29mm RB, 2/pk 16.3 -0.4 ● ● 30314906 Rack, 3x 50ml D29mm FA, 2/pk 16.7 0.0 ● ● 30314907 Rack, 1x 50ml D34mm RB, 2/pk 15.8 -0.9 ● ● 30314908 Rack, 2x 50ml D29mm Rim, 2/pk 15.8 -0.9 ● ● 30314909 Rack, 1x 50ml D29mm Rim, 2/pk 16.1 -0.6 ● ● 30314910 Rack, 5x 25ml D24.5mm RB, 2/pk 16.1 -0.6 ● ● 30314911 Rack, 3x 25ml D24.5mm Rim, 2/pk 15.5 -1.2 ● ● 30314912 Rack, 9x 15ml D17mm RB, 2/pk 16.3 -0.4 ● ● 30314913 Rack, 7x 15ml D17mm FA, 2/pk 16.4 -0.3 ● ● 30314914 Rack, 7x 15ml D17mm Rim, 2/pk 16.1 -0.6 ● ● 30314915 Rack, 8x 10ml D16mm RB, 2/pk 16.3 -0.4 ● ● 30314916 Rack, 14x 5-7ml D13mm RB, 2/pk 15.9 -0.8 ● ● 30314917 Rack, 10x 1.6-7ml D13mm Vac, 2/pk 15.9 -0.8 ● ● 30304367 Rack, 8x 4-10ml D16mm Vac, 2/pk 16.3 -0.4 ● ● 30304368 Adapter, 9x 1.5/2.0ml D11mm, 2/pk 15.7 -1.0 ● ● Blank NO Bucket/Adapter 14.1 30304369 30314821 Rotor, Angle, 6x 250ml FB, ID ● ● ● 13.2 ● ● ● ● 30304370 Adapter, 5x 10ml D16mm RB, 2/pk 13.3 ● ● ● ● 30304371 Adapter, 4x 15ml D17mm FA, 2/pk 13.0 -1.1 ● ● ● ● 12.5 -1.6 ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● 30304372 Adapter, 2x 30ml D26mm RB, 2/pk 30304373 Adapter, 1x 50ml D29mm RB, 2/pk 12.0 -2.1 30304374 Adapter, 1x 50ml D30mm FA, 2/pk Adapter 15x 1.5/2.0ml D11mm MTBx 2 V1 11.7 -2.4 9.5 13.6 -4.6 -0.5 -0.5 30559413 11.8 Blank Adapter 1x 250ml D60mm CBCx 3 V1 Adapter 1x 175/225ml D61mm FAx 2 V1 NO Bucket/Adapter ● ● ● 30314842 Rack, 1x 100ml D46mm RB, 2/pk 14.1 -0.7 ● ● ● 30314843 Rack, 1x 85ml D38mm RB, 2/pk 14.1 -0.7 ● ● ● 30314844 Rack, 1x 50ml D34mm RB, 2/pk 14.2 -0.6 ● ● ● 30314845 Rack, 1x 50ml D29mm RB, 2/pk 14.2 -0.6 ● ● ● 30314846 Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.7 ● ● ● 30314847 Rack, 1x 50ml D29mm Rim, 2/pk 14.3 -0.5 ● ● ● 30314848 Rack, 1x 30ml D25mm Rim, 2/pk 13.8 -1.0 ● ● ● 30314849 Rack, 1x 25ml D25mm RB, 2/pk 14.4 -0.4 ● ● ● 30314850 Rack, 2x 15ml D17mm FA, 2/pk 14.8 0.0 ● ● ● 30559414 Rotor, Sw ing out, 4x 100ml, ID Adapter, 8x 1.5ml D11mm, 2/pk ● -0.9 -0.8 30559412 30314822 Correction (cm) FC5816R Rotor, Sw ing out, 4x 250ml, ID Radius (cm) FC5816 30314820 Description du godet/adaptateur FC5718R 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable Godet/Ada ptateur FC5718 Description du rotor FC5714 Rotor 12.8 -2.3 14.8 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314851 Rack, 1x 15ml D17mm Rim, 2/pk 14.5 -0.3 ● ● ● 30314852 Rack, 7x 15ml D17mm RB, 2/pk 14.5 -0.3 ● ● ● 30314853 Rack, 9x 5-7ml D13mm RB, 2/pk 14.2 -0.6 ● ● ● 30314854 Rack, 10x 1.5ml D11mm, 2/pk 9.9 -4.9 ● ● ● 30314855 Rack, 9x 4.5ml D12mm FB, 2/pk 14.0 -0.8 ● ● ● 30314856 Rack, 7x 5-7ml D13mm RB, 2/pk 11.6 -3.2 ● ● ● 30314857 Rack, 7x 5-7ml D13mm Vac, 2/pk 14.2 -0.6 ● ● ● 30314858 Rack, 5x 4-10ml D16mm Sar, 2/pk 14.3 -0.5 ● ● ● Blank 14.6 ● ● ● 30314861 NO Bucket/Adapter Bucket, 100ml w o Cap, Sealable, 2/pk Bucket, 100ml w / Cap, Sealable, 2/pk 30314864 Rack, 1x 100ml D40mm RB, 2/pk 30314865 30314860 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 14.2 -0.4 ● ● ● 14.2 -0.4 ● ● ● 13.8 -0.8 ● ● ● Rack, 1x 85ml D38mm RB, 2/pk 14.0 -0.6 ● ● ● 30314866 Rack, 1x 50ml D34mm RB, 2/pk 14.0 -0.6 ● ● ● 30314867 Rack, 1x 50ml D29mm RB, 2/pk 14.0 -0.6 ● ● ● 30314868 Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.5 ● ● ● 30314869 Rack, 1x 50ml D29mm Rim, 2/pk 13.9 -0.7 ● ● ● 30314870 Rack, 1x 30ml D25mm Rim, 2/pk 13.1 -1.5 ● ● ● 30314871 Rack, 1x 25ml D25mm RB, 2/pk 14.0 -0.6 ● ● ● 30314872 Rack, 1x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314873 Rack, 2x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314874 Rack, 2x 15ml D17mm Rim, 2/pk 14.0 -0.6 ● ● ● 30314875 Rack, 4x 15ml D17mm RB, 2/pk 14.0 -0.6 ● ● ● 30314876 Rack, 7x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314877 Rack, 5x 1.5ml D11mm, 2/pk 14.0 -0.6 ● ● ● 30314878 Rack, 3x 10ml D17mm Sar, 2/pk 14.0 -0.6 ● ● ● 30314879 Rack, 4x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314880 14.0 -0.6 ● ● ● 14.6 0.0 ● ● ● 14.6 0.0 ● ● ● 30314883 Rack, 4x 10ml D17mm Vac, 2/pk Bucket, 2x 50ml D29mm FA, w / Rack, 2/pk Bucket, 3x 15ml D17mm FA, w / Rack, 2/pk Cy to, Rectangular Bucket, 2/pk 6.3 -8.3 ● ● ● Blank NO Bucket/Adapter 12.0 ● ● ● ● ● ● ● 30314891 Adapter, 48x 1.5/2.0ml D11mm, 2/pk 12.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 10.3 ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 9.6 -0.7 ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 9.6 -0.7 ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 10.0 -0.3 ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 9.8 -0.5 ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 9.3 -1.0 ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 10.3 0.0 ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 9.5 -0.8 ● ● Blank NO Bucket/Adapter 11.5 ● ● ● ● ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 10.9 -0.6 ● ● ● ● ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 11.1 -0.4 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314881 30314882 30314824 30314825 30314826 Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID FR-43 0.0 FR-44 30314827 30314828 30314829 30314830 30314831 30314832 30314834 30314835 30314836 30314838 30314839 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Rotor, Angle, 4x 85ml RB, ID, Hi Rotor, Sw ing out, 16x 50ml, ID Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable Rotor, Angle, 24x 1.5/2.0ml, ID Rotor, Angle, 12x 1.5/2.0ml, ID Blank NO Bucket/Adapter 9.2 ● ● ● ● ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 8.5 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 8.4 -0.7 ● ● ● ● ● ● -0.8 ● ● ● ● ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 8.9 -0.3 ● ● ● ● ● ● 30314896 30314897 Adapter, 1x 50ml D29mm FA, 2/pk 8.6 -0.6 ● ● ● ● ● ● Adapter, 1x 15ml D17mm FA, 2/pk 8.3 -0.9 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 7.5 -1.7 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 8.3 -0.9 ● ● ● ● ● ● Blank NO Bucket/Adapter 16.5 ● ● 30304375 Rack, 4x 50ml D29mm FA, 2/pk 16.5 0.0 ● ● 30314583 Rack, 10x 15ml D17mm FA, 2/pk 16.5 0.0 Blank NO Bucket/Adapter 13.0 30130894 Adapter, 1x 15ml D17mm RB, 2/pk 12.8 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk Blank NO Bucket/Adapter 11.0 30130891 Adapter, 1x 30ml D26mm, 2/pk 10.7 -0.3 30130892 Adapter, 1x 16ml D18mm, 2/pk 10.3 30130893 Adapter, 1x 15ml D17mm, 2/pk 10.6 30130894 Adapter, 1x 15ml D17mm RB, 2/pk 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 30130886 ● ● ● ● ● ● ● ● -0.2 ● ● ● ● ● ● 12.2 -0.8 ● ● ● ● ● ● 10.4 -2.6 ● ● ● ● ● ● 8.9 -4.1 ● ● ● ● ● ● ● ● ● ● ● ● -0.7 ● ● ● -0.4 ● ● ● 10.6 -0.4 ● ● ● 10.2 -0.8 ● ● ● 8.3 -2.7 ● ● ● Adapter, 1x 1.5/2.0ml D11mm, 6/pk 6.7 -4.3 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130891 Adapter, 1x 30ml D26mm, 2/pk 8.2 -0.2 ● ● ● ● ● ● 30130892 Adapter, 1x 16ml D18mm, 2/pk 7.9 -0.5 ● ● ● ● ● ● 30130893 Adapter, 1x 15ml D17mm, 2/pk 8.0 -0.4 ● ● ● ● ● ● 30314892 Adapter, 1x 15ml D17.5mm FA, 2/pk 7.7 -0.7 ● ● ● ● ● ● Blank NO Bucket/Adapter 12.5 ● ● ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 12.2 -0.3 ● ● ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 10.5 -2.0 ● ● ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 9.0 -3.5 ● ● ● ● ● Blank NO Bucket/Adapter 11.0 ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 10.6 -0.4 ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 9.1 -1.9 ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 7.7 -3.4 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.2 ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.7 -0.7 ● ● Blank NO Bucket/Adapter 9.4 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 9.1 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 8.4 Blank NO Bucket/Adapter 8.5 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 Blank NO Bucket/Adapter 6.5 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 6.3 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 5.6 ● ● ● ● ● ● ● ● ● -0.3 ● ● ● ● ● -1.0 ● ● ● ● ● ● ● ● ● ● ● ● -0.3 ● ● ● ● ● ● ● -1.0 ● ● ● ● ● ● ● ● ● -0.2 ● ● -0.9 ● ● Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Blank Adapter, 1x 1.5ml for ov er16000x g, 6/pk NO Bucket/Adapter 19.3 ● ● 30553138 Adapter 32x 7ml D13mm RBx 2 V1 18.7 -0.6 ● ● 30553135 Adapter 18x 15ml D17mm RBx 2 V1 18.8 -0.5 ● ● 30553132 Adapter 8x 30ml D26mm RBx 2 V1 19.1 -0.2 ● ● 30553131 Adapter 7x 50ml D29mm RBx 2 V1 19.1 -0.2 ● ● 30553127 Adapter 3x 100ml D41mm RBx 2 V1 19.1 -0.2 ● ● 30553125 18.9 -0.4 ● ● 19.1 -0.2 ● ● 30553128 Adapter 1x 250ml D62mm FBx 2 V1 Adapter 1x 225/190ml D61mm FAx 2 V1 Adapter 5x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553130 Adapter 7x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553133 Adapter 14x 15ml D17mm FAx 2 V1 19.2 -0.1 ● ● 30553134 Adapter 17x 15ml D17mm FAx 2 V1 Adapter 18x 5/10ml D16mm Vac/Sarx 2 V1 Adapter 38x 1.1/1.4ml D8mm Sarx 2 V1 19.0 -0.3 ● ● 19.0 -0.3 ● ● 18.9 -0.4 ● ● Adapter 21x 10ml D16mm RBx 2 V1 Adapter 27x 1.8/7ml D13mm Vac/Sarx 2 V1 NO Bucket/Adapter 18.9 -0.4 ● ● 18.8 -0.5 ● ● ● ● 13.7 -0.8 ● ● 13.4 -1.1 ● ● 12.4 -2.1 ● ● 12.6 -1.9 ● ● 30559421 Adapter 1x 250ml D61.5mm FBx 2 V1 Adapter 1x 175/225ml D61.5mm FAx 2 V1 Adapter 1x 50ml D29.5mm FAx 2 V1 Adapter 1x 250ml D60.5mm CBCx 2 V1 Adapter 4x 30ml D25.5mm RBx 2 V1 13.8 -0.7 ● ● 30559420 Adapter 3x 50ml D29mm RBx 2 V1 14.3 -0.2 ● ● 30559422 Adapter 6x 15ml D17mm FAx 2 V1 13.8 -0.7 ● ● 30314900 30553085 Rotor Sw ingOut 4x 750ml D99mm ID SL 30553126 30553140 30559377 30553136 30553139 Blank 30564850 30559417 30553086 Rotor Angle 4x 500ml D69mm ID FR-45 30559419 30559416 6.4 -0.1 14.5 ● ●  FR-46 Centrifugeuse FrontierTM FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.7 Tableau 7 : Formulaire de rachat/Certificat de décontamination Compléter le formulaire en lettres capitales Ce formulaire doit accompagner tous les retours d'équipements et d'ensembles ! La déclaration complète sur la décontamination est un prérequis avant tout traitement du produit retourné. Si aucune explicat ion correspondante n'est incluse, il sera procédé à la décontamination aux frais du cl ient. Prénom ; nom :_______________________________________ Société/Entreprise :___________________________________ Ville : ___________________________________ ___________ CODE POSTAL : _____________ lieu :____________________ Téléphone : _____________ Fax :____________________ E-Mail : _____________________________________ Pos. Crowd Decontaminated object Serial number Description / Comment 1 2 3 4 Les pièces répertoriées ci-dessus sont-elles en contact avec les substances suivantes ? Solutions aqueuses, tampons, acides, al calis dangereux pour la santé :...............  Oui  Non Agents infectieux potentiels : .................................. ...............................................  Oui  Non Réactifs organiques et solvants : ……………………………………………………..  Oui  Non Substances radioactives :............................................................. α.. β.. γ...  Oui  Non Protéines dangereuses pour la santé :………………………………………………..  Oui  Non ADN : …………………………………………………………………………………..…  Oui  Non Ces substances sont-elles entrées en contact avec l'équipement/l'assemblage ?  Oui  Non Laquelle, si oui : ________________________________________________________________ Description des mesures de décontamination des pièces répertoriées : Je confirme que la décontamination est adéquate : Entreprise/Service ________________________ Lieu et date : ________________________ Signature de la personne autorisée :_____________________________________ Skizzenkarte von FC5714 / FC5718 / FC5816 / FC5916 1 2 7 4 3 6 5 Abbildung.1 Skizzenkarte von FC5718R 1 2 7 3 4 6 5 Abbildung.2 Skizzenkarte von FC5816R / FC5916R 1 2 7 3 6 4 5 Abbildung.3 1 Zentrifugendeckel 3 Anzeige 5 Hauptstromschalter 7 Notfreigabe 2 Rotorfenster 4 Funktionstafel 6 Stromanschluss Funktionsetikett für FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Abbildung.4 1 4 5 6 7 8 2 3 9 10 11 12 Abbildung.5 1. 3. 5. LCD-Anzeige Zentrifugation starten Beschleunigungs/Verlangsamungsintensitätsmo dell (Acc/Dec) und Auswahl 7. Zeiteinstellungsmodell 9. Kurze/Schnelle Drehzentrifugation 11. Zentrifugation stoppen / Einstellung 2. 4. Programmeinstellungsmodell RPM/RFC-Modell und Auswahl 6. Temperatureinstellungsmodell (Nur F5515R) 8. Speichereinstellungsmodell 10. Freigabedeckel 12. Anpassungsknopf/Wählscheibe: Nummer ändern LCD-Anzeige Das folgende Bild zeigt die einzelnen Elemente der LCD-Anzeige. Abbildung.6 M2 M3 M1 M12 M15 M13 M4 M9 M11 M10 M5 M6 M7 A-1 M8 A-2 M14 A-3 A-4 Abbildung.7 Anzeigefelder: A-1 A-2 A-3 A-4 Anzeigefeld – "RPM/RCF" Anzeigefeld – "Acc/Dec" "Service" Anzeigefeld – "Time/Prog" Anzeigefeld – "Temp" Mitteilungen/Logos der Anzeigefelder M1 M4 M7 M10 M13 "close" "Rotor-Nr." "accel" "program" "h m s" M2 M5 M8 M11 "open" "rpm" "decel" "error" M3 M6 M9 M12 M14 "rotor" "rcf" "radius" "service" "temperature" M15 "precool" Rotor informations tabelle 21 FC5916R 30314820 FC5916 20 FC5816R Rotor, Angle, 24x1.5/2.0ml, ID, Sealable FC5816 30304361 FC5718R 61 Kapazitä t FC5718 Bestell-Nr. FC5714 Kompatibel Rotor Nr. Anzeige ● ● ● ● ● ● ● Rotor, Swing out, 4x250ml, ID ● ● 30314821 Rotor, Angle, 6x250ml FB, ID ● 22 30314822 Rotor, Swing out, 4x100ml, ID ● ● ● 23 30314823 Rotor, Swing out, 4x100ml, ID, Sealable ● ● ● 24 30314824 Rotor, Swing out, 2x3MTP, ID ● ● ● 25 30314825 Rotor, Angle, 6x85ml RB, ID, Hi ● ● 26 30314826 Rotor, Angle, 6x85ml RB, ID ● 27 30314827 Rotor, Angle, 4x85ml RB, ID, Hi ● 28 30314828 Rotor, Swing out, 16x50ml, ID 29 30314829 Rotor, Angle, 10x50ml FA, ID 30 30314830 Rotor, Angle, 6x50ml RB/FA, ID 31 30314831 Rotor, Angle, 6x50ml RB, ID, Hi 32 30314832 Rotor, Angle, 30x15ml RB/FA, ID 33 30314833 Rotor, Angle, 20x10ml RB, ID, Hi 34 30314834 Rotor, Angle, 12x15ml RB/FA, ID 35 30314835 Rotor, Angle, 44x1.5/2.0ml, ID 36 30314836 Rotor, Angle, 30x1.5/2.0ml, ID, Sealable 38 30314838 Rotor, Angle, 24x1.5/2.0ml, ID 39 30314839 40 ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● Rotor, Angle, 12x1.5/2.0ml, ID ● ● 30314840 Rotor, Angle, 64x0.5ml, ID ● ● 41 30314841 Rotor, Angle, 4x8-w PCR Strip, ID ● ● ● ● 85 30553085 Rotor SwingOut 4x750ml D99mm ID SL ● ● 86 30553086 Rotor Angle 4x500ml D69mm ID ● ● ● ● ● ● ● INHALTSANGABE 1. EINLEITUNG ............................................................................................................ ..............................................................1 1.1 BESCHREIBUNG UND VERWENDUNGSZWECK ....................................................................................................................1 1.2 KURZBESCHREIBUNG ...........................................................................................................................................................1 1.3 DEFINITION DER SIGNALWARNUNGEN UND SYMBOLE ......................................................................................................1 1.4 SICHERHEITSVORKEHRUNGE N ............................................................................................................................................2 1.4.1 BENUTZER ....................................................................................................................................................................2 1.4.2 ROTOR UND ZUBEHÖ RTEILE ........................................................................................................................................2 1.4.3 MAßNAHMEN ZU IHREM EIGENEN SCHUTZ ................................................................................................................2 1.4.4 SCHLIEßEN SIE DIE FOLGENDEN UMWELTEINFLÜ SSE AUS ..........................................................................................2 1.4.5 MAßNAHMEN FÜ R DIE BETRIEBSSICHERHEIT..............................................................................................................3 1.4.6 GEFAHR UND VORKEHRUNGE N ...................................................................................................................................3 1.4.7 ABKÜ RZUNGEN, DIE IN DIESEM HANDBUCH BENUTZT WERDEN................................................................................3 2. INSTALLATION .......................................................................................................................................................................4 2.1 AUSPACKEN .........................................................................................................................................................................4 2.1.1 LIEFERPAKET ................................................................................................................................................................4 2.2 AUSWAHL DES STANDORTS.................................................................................................................................................4 2.3 INSTALLATION......................................................................................................................................................................4 2.4 SICHERHEITSVORKEHRUNGE N WÄHREND DES BETRIEBS UND GARANTIE .........................................................................5 3. BETRIEB ............................................................................................................................................................... ..................5 3.1 BEFESTIGEN UND LADEN DES ROTORS ................................................................................................................................5 3.1.1 INSTALLATION VON ROTOREN.....................................................................................................................................5 3.1.2 BELADEN VON WINKLIGEN ROTOREN .........................................................................................................................6 3.1.3 BELADEN VON AUSSCHWE NKROTO RE N ......................................................................................................................6 3.1.4 BELADEN UND Ü BERLADEN DER ROTOREN .................................................................................................................7 3.1.5 ENTFERNUNG DES ROTORS .........................................................................................................................................7 3.2 KONTROLLE DES DECKELS ....................................................................................................................................................7 3.2.1 DECKEL Ö FFNEN ...........................................................................................................................................................7 3.2.2 DECKELVERSCHLUSS ....................................................................................................................................................8 3.3 VORAUSWAHL .....................................................................................................................................................................8 3.3.1 VORAUSWAHL DER GESCHWINDIGKEIT / RCF-WERT ..................................................................................................8 3.3.2 VORAUSWAHL DER LAUFZEIT ......................................................................................................................................9 3.3.3 VORAUSWAHL DER BREMSINTENSITÄT UND BESCHLEUNIGUNG ...............................................................................9 3.3.4.VORAUSWAHL DER TEMPERATUR (NUR GEKÜ HLTE MODELLE)................................................................................10 3.3.5 VORKÜ HLUNG (NUR GEKÜ HLTE MODELLE)...............................................................................................................10 3.4 RADIUSKORREKTUR ...........................................................................................................................................................10 3.5 PROGRAMM ......................................................................................................................................................................11 3.5.1 SPEICHERUNG VON PROGRAMMEN ..........................................................................................................................11 3.5.2 AUFRUF VON GESPEICHERTEN PROGRAMMEN ........................................................................................................12 3.5.3 VERLASSEN DES PROGRAMMMODUS .......................................................................................................................12 3.6 STARTEN UND STOPPEN DER ZENTRIFUGE .......................................................................................................................12 3.6.1 STARTEN DER ZENTRIFUGE ........................................................................................................................................12 3.6.2 DIE "STOP"-TASTE ......................................................................................................................................................13 3.7 FESTSTELLUNG EINES UNGLEICHGEWICHTS......................................................................................................................13 4. EINSTELLUNG ......................................................................................................................................................................1 4 4.1 GRUNDANPA SSUNGE N ......................................................................................................................................................14 4.1.1 ZUGRIFF AUF DEN MODUS "BETRIEBSDATEN" ..........................................................................................................14 4.1.2 TEMPERATURANGA BE ...............................................................................................................................................15 4.1.3 SIGNAL EIN-/AUSSCHALTE N .......................................................................................................................................15 4.1.4 LAUTSTÄ RKE-VORAUSWAHL DES TONSIGNALS .........................................................................................................16 4.1.5 LIEDAUSWAHL FÜ R DAS TONSIGNAL AM ENDE DES DURCHLAUFS ..........................................................................16 4.1.6 TASTATURGERÄUSCH EIN-/ AUSSCHALTEN ...............................................................................................................16 4.1.7 BETRIEBSDATEN AUFRUFEN ......................................................................................................................................17 5. WARTUNG............................................................................................... ............................................................................1 8 5.1 WARTUNG UND REINIGUNG .............................................................................................................................................18 5.1.1 ALLGEMEIN ...............................................................................................................................................................18 5.1.2 REINIGUNG UND DESINFEKTION DER EINHEIT ..........................................................................................................18 5.1.3 REINIGUNG UND DESINFEKTION DES ROTORS ..........................................................................................................19 5.1.4 DESINFEKTION VON ALUMINIUMROTORE N ..............................................................................................................19 5.1.5 DESINFEKTION DER PP-ROTOREN ..............................................................................................................................19 5.1.6 GLASBRUCH ...............................................................................................................................................................19 5.2 LEBENSZEIT DER ROTOREN, BEHÄLTER, ZUBEHÖ RTEILE ...................................................................................................19 6. PROBLEMBEHEBUNG ..........................................................................................................................................................20 6.1 FEHLERMITTEILUNG: URSACHE / LÖ SUNG ........................................................................................................................20 6.2 UNTERSUCHUNG DER MÖ GLICHEN FEHLERMITTEILUNGEN UND IHRER LÖ SUNGEN.......................................................20 6.2.1 DECKELFREIGABE BEI STROMAUSFALL (NOTFALLDECKELFREIGABE) ........................................................................20 6.2.2 BESCHREIBUNG DES FEHLERMITTEILUNGSSYSTE M S .................................................................................................21 7. ENTGEGENNAHME VON ZU REPARIERENDEN ZENTRIFUGEN ............................................................................................21 8. TRANSPORT, LAGERUNG UND ENTSORGUNG ....................................................................................................................22 8.1 TRANSPORT .......................................................................................................................................................................22 8.2 LAGERUNG .........................................................................................................................................................................22 8.3 TRANSPORT, INSTALLATION, Ü BERTRAGUNG UND ENTSORGUNG DER ZENTRIFUGE FC5714/FC571 8/FC57 18 R/FC5 816 /FC5 816 R ..........................................................................................................................22 8.3.1 TRANSPORT................................................................................................................................................................22 8.3.2 INSTALLATION ............................................................................................................................................................22 8.3.3 VERPACKUNG.............................................................................................................................................................23 8.3.4 WEITERGABE DES GERÄ TS .........................................................................................................................................23 9. TECHNISCHE DATEN............................................................................................................................................................24 9.1 SPEZIFIKATIONEN ..............................................................................................................................................................24 9.1.1 ZENTRIFUGE FC5714 ..................................................................................................................................................24 9.1.2 ZENTRIFUGE FC5718 ..................................................................................................................................................25 9.1.3 ZENTRIFUGE FC5718R ................................................................................................................................................26 9.1.4 CENTRIFUGE FC5816 ..................................................................................................................................................27 9.1.5 ZENTRIFUGE FC5816R ................................................................................................................................................28 9.2 ZEICHNUNGEN UND AUSMAßE .........................................................................................................................................29 10. COMPLIANCE ......................................................................................................................................................................31 11. ANHANG .............................................................................................................................................................................32 11.ϭ TABELLE ϭ: ZULÄSSIGES NETTOGEWICHT ........................................................................................................................33 11.Ϯ TABELLE Ϯ: NIEDRIGSTE TEMPERATUREN BEI MAX.-GESCHWINDIGKEIT (NUR GEKÜ HLTE MODELLE)...........................34 11.ϯ TABELLE ϯ: MAX. GESCHWINDIGKEIT UND RCF-WERTE FÜ R ZULÄSSIGE ROTOREN .......................................................͘35 11.ϰ TABELLE ϰ: BESCHLEUNIGUNGS- UND VERLANGSAMUNGSZEIT E N ................................................................................37 11.ϱ TABELLE ϱ: FEHLERMITTEILUNGEN .................................................................................................................................40 11.ϲ TABELLE ϲ: RADIUSKORREKTUR UND ADAPTERSPEZIFIKATIONEN..................................................................................41 11.ϳ TABELLE ϳ: RÜ CKNAHMEFORMULAR / DEKONTAMINIERUNGSZERTIFIKAT....................................................................45 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-1 1. EINLEITUNG 1.1 Beschreibung und Verwendungszweck Vielen Dank dafür, dass Sie ein OHAUS-Produkt ausgewählt haben. Alle Symbole stellen Sicherheitsanweisungen dar und weisen auf potentielle Gefahrensituationen hin. Bitte lesen Sie das Handbuch vollständig durch, bevor Sie die Frontier TM MultiPro-Zentrifugen verwenden, um Fehlbedienungen zu vermeiden. FrontierTM MultiPro Zentrifugen wurden für die Trennung von Materialien oder Gemischen mit unterschiedlichen Dichten entwickelt. OHAUS-Zentrifugen sind ausschließlich für die Innenanwendung und für die Anwendung durch qualifiziertes Personal vorgesehen. 1.2 Kurzbeschreibung Die Modelle FC5714/FC5718/FC5816/FC5916 sind nicht gekühlte Universalzentrifugen. Die Modelle FC5718R/FC5816R/FC5916R sind gekühlte Universalzentrifugen. Alle drei Modelle werden in zwei Spannungsvarianten, 230V oder 120V angeboten. Die Zentrifugen können mit Schwenk- und Winkelrotoren verwendet werden. Alle Parameter sind über Tasten zugänglich und werden mit dem zentralen Einsteller ausgewählt. Alle vorgewählten und aktuellen Werte werden auf der LCD-Anzeige dauerhaft angezeigt. Die Zentrifuge wird durch einen wartungsfreien Induktionsmotor angetrieben. Detaillierte technische Daten finden Sie im Abschnitt "Technische Daten". 1.3 Definition der Signalwarnungen und Symbole Sicherheitsanmerkungen sind mit Signalwörtern und Warnsymbolen gekennzeichnet. Diese zeigen Sicherheitsproblem und Warnungen an. Eine Missachtung der Sicherheitsanmerkungen kann zu Verletzungen an Personen, Schäden an Instrumenten, sowie Fehlfunktionen und falschen Ergebnissen führen. Der Grad der Gefahr ist ein Teil einer Sicherheitsanmerkung und unterscheidet die möglichen Ergebnisse ihrer Nichteinhaltung. Signalwörter GEFAHR WARNUNG VORSICHT ACHTUNG ANMERKUNG Wird zu schweren Verletzungen oder zum Tode führen, falls nicht vermieden. Für eine gefährliche Situation mit mittlerem Risiko, die möglicherweise zu Verletzungen oder zum Tode führt, falls nicht vermieden. Für eine gefährliche Situation mit geringem Risiko, die du eine Schaden am Gerät oder Eigentum oder zu einem Datenverlust oder zu Verletzungen führt, falls nicht vermieden. Für wichtige Informationen über das Produkt Kann zu Schäden an der Ausrüstung führen, falls nicht vermieden Für nützliche Informationen über das Produkt Warnsymbole Allgemeine Gefahr Stromschlaggefahr Wechselstrom Biologische Gefahr Explosion Zerkleinerung DE-2 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Warn- und Informationsschilder auf der Zentrifugenoberfläche Es müssen stets vier Trägers an vier Stellen bei Ausschwenkrotoren verwendet werden, da ansonsten Beschädigungen an der Zentrifuge auftreten können. Eine derartige Beschädigung wird von der Produktgarantie nicht abgedeckt. Achtung! Prüfen Sie die Befestigung der Rotormutter vor jedem Lauf. Ziehen Sie den Netzstecker, bevor Sie das Gehäuse oder die Notfreigabe öffnen. Drehrichtung - Drehung im Uhrzeigersinn für den Rotorantrieb Referenz zum Laden von Rotoren 1.4 Sicherheitsvorkehrungen 1.4.1 Benutzer OHAUS-Zentrifugen sind ausschließlich für die Innenanwendung und für die Anwendung durch qualifiziertes Personal vorgesehen. Das Gerät kann nur von einem geschulten Fachmitarbeiter betrieben werden. Sie müssen das Betriebshandbuch gründlich durchlesen und sich mit der Funktion des Gerätes vertraut machen. 1.4.2 Rotor und Zubehörteile Es sollten nur Originalrotoren und Zubehörteile von OHAUS verwendet werden. Jede andere Nutzung oder beabsichtigte Nutzung wird als ungeeignet erachtet. OHAUS haftet nicht für Schäden, die aus einer ungeeigneten Anwendung resultieren. VORSICHT: Lesen Sie alle Sicherheitswarnungen durch, bevor Sie diese Ausrüstung installieren, anschließen oder warten. Ein Versäumnis, diese Warnungen zu befolgen, könnte zu Verletzungen an Personen und/oder Beschädigungen am Eigentum führen. Bewahren Sie die Anweisungen zwecks zukünftiger Bezugnahme auf. 1.4.3 Maßnahmen zu Ihrem eigenen Schutz WARNUNG: Arbeiten Sie niemals in einer explosionsgefährdeten Umgebung! Das Gehäuse des Instruments ist nicht gasdicht. (Explosionsgefahr aufgrund von Funkenbildung, Korrosion hervorgerufen durch das Eindringen von Gasen) WARNUNG: Befolgen Sie, wenn Sie Chemikalien und Lösungsmittel benutzen, die Anweisungen des Herstellers und die allgemeinen Regeln zur Laborsicherheit. WARNUNG: Die Zentrifuge ist nicht versiegelt. Benutzen Sie geeignete Schutzmaßnahmen, wenn Sie die Zentrifuge für infektiöse und pathogene Proben. Befolgen Sie angemessene Sicherheitsmaßnahmen, wenn Sie mit diesen Proben arbeiten. 1.4.4 Schließen Sie die folgenden Umwelteinflüsse aus       Starke Vibrationen Direktes Sonnenlicht Atmosphärische Feuchtigkeit grösser als 80% Korrosive Gase vorhanden Temperaturen unter 2 °C und über 35 °C Starke elektrische oder magnetische Felder: FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-3 WARNUNG: Innerhalb des Gehäuses besteht die Gefahr von elektrischen Stromschlägen. Das Gehäuse sollten nur von autorisiertem und qualifiziertem Personal geöffnet werden. Entfernen Sie vor dem Öffnen alle Stromverbindungen zur Einheit. 1.4.5 Maßnahmen für die Betriebssicherheit  Entschrauben Sie nicht die beiden Hälften des Gehäuses.  Trocknen Sie jede ausgelaufene Flüssigkeit sofort ab! Das Instrument ist nicht wasserdicht.  Prüfen Sie, dass die Eingangsspannung und die Steckerart des Geräts mit der lokalen Stromversorgung vereinbar sind.  Stecken Sie das Stromkabel nur in einen ordnungsgemäß geerdeten Stromempfänger.  Benutzen Sie nur ein Stromkabel mit einem Nennwert, der die Spezifikationen auf dem Ausrüstungsetikett überschreitet.  Platzieren Sie die Ausrüstung nicht so, dass es schwierig ist, das Stromkabel vom Stromempfänger zu trennen.  Stellen Sie sicher, dass das Stromkabel kein potentielles Hindernis oder eine Stolperfalle darstellt.  Die Ausrüstung ist nur für den Innengebrauch vorgesehen. Benutzen Sie die Ausrüstung nur in trockenen Umgebungen.  Benutzen Sie nur angemessene Zubehörteile.  Betreiben Sie die Ausrüstung nur unter Umgebungsbedingungen, die in diesen Anweisungen festgelegt werden.  Trennen Sie die Ausrüstung bei der Reinigung von der Stromversorgung.  Betreiben Sie die Ausrüstung nicht in gefährlichen oder unstabilen Umgebungen.  Instandhaltungsarbeiten sollten nur von autorisiertem Personal durchgeführt werden. 1.4.6 Gefahr und Vorkehrungen Um Mensch und Umwelt zu schützen, sollten die folgenden Vorkehrungen eingehalten werden:  Während der Zentrifugation ist der Aufenthalt von Personen und die Aufstellung von gefährlichen Materialien innerhalb von 30 cm um die Zentrifuge herum, gemäß den Bestimmungen von EN 61010-2-020. verboten.  FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R sind nicht explosionssicher und darf daher nicht in expliosionsgefährdeten Bereichen oder Standorten betrieben werden. Die Zentrifugation von brennbaren, explosiven, radioaktiven oder solchen Substanzen, die chemisch mit hoher Energie reagieren, ist strengstens verboten. Die endgültige Entscheidung in Bezug auf die Risiken in Verbindung mit der Verwendung von solchen Substanzen liegt in der Verantwortung der Benutzer der Zentrifuge.  Drehen Sie niemals toxisches oder pathogenes Material ohne angemessene Sicherheitsvorkehrungen, d.h. Zentrifugation von Kübeln, / Rohren mit fehlender oder beschädigter hermetischer Versiegelung ist strengstens verboten. Der Benutzer ist verpflichtet, ein angemessenes Desinfektionsverfahren durchzuführen, für den Fall, dass gefährliche Substanzen die Zentrifuge und/oder ihre Zubehörteile kontaminiert haben. Wenn infektiöse Substanzen zentrifugiert werden, achten Sie stets auf die allgemeinen Laborvorkehrungen. Falls erforderlich, setzen Sie sich mit Ihrem Sicherheitsbeauftragten in Verbindung!  Es ist verboten, die Zentrifuge mit Rotoren zu betreiben, die nicht für diese Einheit aufgeführt sind.  Öffnen Sie den Deckel der Zentrifuge unter keinen Umständen, während der Rotor noch läuft oder sich mit einer Geschwindigkeit von> 2m / s dreht 1.4.7 Abkürzungen, die in diesem Handbuch benutzt werden Symbol/Abkürzungen RPM RCF Einheit -1 [min ] rpm Beschreibung [x g] relative Zentrifugalkraft PCR Umdrehungen pro Minute Polymerasekettenreaktion PP - Polypropylen PC - Polykarbonat accel - Beschleunigung decel - Verlangsamung prog - Programm DE-4 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 2. INSTALLATION 2.1 Auspacken Entnehmen Sie Ihre Zentrifuge und jede ihrer Komponenten vorsichtig aus der Verpackung. Die enthaltenen Komponenten variieren je nach Zentrifugenmodell (siehe Tabelle unten). Sichern Sie die Verpackung, um eine sichere Lagerung und Transport zu gewährleisten. Die Bedienungsanleitung muss immer mit der Zentrifuge aufbewahrt werden! Rotor(en) / Zubehörteile werden separat verpackt. WARNUNG: Hebegefahr. Das Heben durch eine einzige Person kann Unfälle verursachen. Benutzen Sie ein mechanisches Hebegerät oder verknüpfen Sie Hebeverfahren, wenn Sie die Ausrüstung heben oder bewegen. Siehe Abschnitt 8.3 für Details über das Anheben des Gerätes 2.1.1 Lieferpaket Menge 1 1 1 1 1 Beschreib u ng Zentrifuge FC5714/F C5718/F C5718R/F C5816/F C5816R/FC5916/FC5916R Stromkabel G ar anti ek ar te Betriebs handbuc h/Kurzanleitung Rotors c hl üss el 2.2 Auswahl des Standorts ANMERKUNG Vermeiden Sie übermäßige Vibrationen, Wärmequellen, Luftstrom oder schnelle Temperaturveränderungen.    Die Zentrifuge sollte auf einer ebenen und festen Oberfläche installiert werden, möglichst auf einem Laborschrank / Tisch oder einer anderen festen, vibrationsfreien Oberfläche. Während der Zentrifugation muss die Zentrifuge so platziert werden, dass auf jeder Seite des Gerätes ein Mindestabstand von 30 cm entsprechend den Normen EN 61010-2-020 vorhanden ist. Stellen Sie die Zentrifuge nicht neben ein Fenster oder eine Heizung, wo sie einer übermäßigen Hitze ausgesetzt sein könnte, da die Leistung des Gerätes auf einer Umgebungstemperatur von 23 ° C basiert. 2.3 Installation Befolgen Sie die folgenden Schritte:  Prüfen Sie, ob die Stromversorgung mit den auf dem Typenschild des Herstellers angegebenen Werten übereinstimmt.  Für FC5714/FC5718/FC581 sollte die Stromleitung durch einen Leistungsschalter von 10 A (Typ K) geschützt werden.  Für FC5718R/FC5816R /FC5916/FC5916R sollte die Stromleitung durch einen Leistungsschalter von 16 A (Typ K) geschützt werden.       Im Notfall muss ein Notausschalter außerhalb des Raumes installiert werden, um die Stromversorgung vom Gerät zu trennen. Verbinden Sie die Zentrifuge mit einer geerdeten Steckdose. Schließen Sie die Zentrifuge ans Netz an. (Die Steckdose für das Stromkabel muss für die Trennung leicht zugänglich sein) Schalten Sie das Instrument ein, indem Sie den Hauptstromschalter benutzen. Öffnen Sie den Deckel, indem Sie die Tür öffnen-Taste verwenden. Entfernen Sie das Transportsicherung vom Motor. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-5 2.4 Sicherheitsvorkehrungen während des Betriebs und Garantie           Betreiben Sie die Zentrifuge nicht, wenn sie nicht korrekt installiert ist. Lehnen Sie sich während des Betriebs nicht an die Zentrifuge. Halten Sie aus betrieblichen Gründen einen Abstand von mehr als 30 cm. Stellen Sie keine potenziell gefährlichen Stoffe innerhalb des Sicherheitsbereichs von 30 cm. Betreiben Sie die Zentrifuge nicht, wenn sie zerlegt wurde (z. B. ohne Gehäuse). Betreiben Sie die Zentrifuge nicht laufen lassen, wenn an mechanischen oder elektrischen Bauteile gearbeitet wird. Verwenden Sie keine Zubehörteile wie Rotoren und Behälter, die nicht exklusiv von der OHAUS Corporation zugelassen werden, außer handelsüblichen Zentrifugenröhrchen aus Glas oder Kunststoff. Drehen Sie keine extrem korrosiven Stoffe, da diese die Materialien beschädigen oder schwächen können. Die Zentrifuge darf nicht mit Rotoren oder Behälter betrieben werden, die Anzeichen auf Korrosion oder mechanischen Beschädigungen aufweisen. Der Hersteller ist für die Sicherheit und Zuverlässigkeit der Zentrifuge verantwortlich, nur wenn: 1) Das Gerät wird gemäß dieser Bedienungsanleitung betrieben. 2) Änderungen, Reparaturen oder sonstige Einstellungen werden von durch OHAUS autorisiertem Personal durchgeführt und die Elektroinstallation entspricht dem entsprechenden Electrical Code. ANMERKUNG Garantie Die Zentrifuge wurde einer gründlichen Prüfung und Qualitätskontrolle unterzogen. Im unwahrscheinlichen Fall von Fertigungsfehlern, werden die Zentrifugen und Rotoren durch eine Garantie abgedeckt. Diese Garantie erlischt bei unsachgemäßer Handlung, Beschädigung und Nachlässigkeit sowie bei unsachgemäßen Ersatzteilen und / oder Zubehör oder unberechtigter Modifikation des Gerätes. Der Hersteller behält sich die technischen Modifikationsrechte hinsichtlich technischer Verbesserungen vor! 3. BETRIEB 3.1 Befestigen und Laden des Rotors 3.1.1 Installation von Rotoren Reinigen Sie die Antriebswelle sowie die Spannzange mit einem sauberen, fettfreien Tuch. Setzen Sie den Rotor auf die Antriebswelle. (Siehe Abbildung unten). Achten Sie darauf, dass der Rotor vollständig auf der Motorwelle montiert ist. Motorwelle und Kammer Abbildung.8 Mutter für Rotor Werkzeug für Rotor mit Mutter Abbildung. 9 Werkzeug für Rotor ohne Mutter Schnappdeckel Abbildung.10 Schraubdeckel DE-6 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Halten Sie den Rotor mit einer Hand fest und befestigen Sie den Rotor an der Welle, indem Sie die Befestigungsmutter im Uhrzeigersinn drehen. Ziehen Sie die Befestigungsmutter mit dem beigefügten Rotorschlüssel fest (Siehe Abbildungen 9-10) Wir werden ein Werkzeug für einen mutterlosen Rotor mit Zentrifuge bereitstellen, das Werkzeug für den Mutterrotor wird mit Rotor bereitgestellt. ACHTUNG! Vergewissern Sie sich, dass die Befestigungsschraube vor jedem Lauf ordnungsgemäß installiert ist. (Siehe Abbildung 9-10) Die Zentrifuge darf nicht mit Rotoren oder Behälter betrieben werden, die Anzeichen auf Korrosion oder mechanischen Beschädigungen aufweisen. Betreiben Sie die Zentrifuge nicht mit extrem korrosiven Substanzen, die den Rotor, die Behälter und Materialien beschädigen könnten. Bei Fragen wenden Sie sich bitte an den Hersteller! 3.1.2 Beladen von winkligen Rotoren Rotoren müssen symmetrisch und mit gleichem Gewicht geladen werden (Siehe Abbildung unten). Dieser Adapter kann nur mit den entsprechenden Gefäßen beladen werden. Die Gewichtsdifferenzen zwischen gefüllten Gefäßen sollten so gering wie möglich gehalten werden. Daher empfehlen wir, die mit einer Waage auszutarieren. Dies reduziert die Abnutzung des Antriebs und den akustischen Betriebslärm Die Maximallast pro Loch wird auf dem Rotor angegeben. Abbildung.11 FALSCH Abbildung.12 KORREKT ( 6 .Röhren) 3.1.3 Beladen von Ausschwenkrotoren Beladen der Behälter / Gefäßen muss in Übereinstimmung mit der unteren Abbildung erfolgen. Sie darf z.B. von einem 4-Place-Rotor mit nur 2 beladenen Behälter betrieben werden. Doch die beladenen Behälter müssen sich gegenüberstehen. Stellen Sie sicher, dass auch die unbeladenen Behälter in den Rotor eingelegt werden (siehe unten) Im Prinzip können Ausschwenkrotoren nicht in Betrieb genommen werden, bis nicht alle Behälter oder Gestelle in den Rotor gestellt sind. Die Schrauben am Rotor müssen mit dem "High TEF Öl" gefettet werden. Die Probenröhrchen müssen nach Augenmaß gleichmäßig gefüllt und in die Bohrungen oder Rohrständer gesteckt werden. Die Gewichtsdifferenz der geladenen Behälter sollte 1,0 g nicht überschreiten. ACHTUNG! Ausschwenkrotoren können nur in Betrieb genommen werden, wenn alle Standorte mit vier Behälter oder vier Trägern gefüllt sind - Mischen Sie keine Behälter und Träger zusammen! ACHTUNG! Die Zentrifuge darf nicht mit Rotoren oder Behälter betrieben werden, die Anzeichen auf Korrosion oder mechanischen Beschädigungen aufweisen. Betreiben Sie die Zentrifuge nicht mit extrem korrosiven Substanzen, die den Rotor und die Behälter beschädigen könnten. Bei Fragen wenden Sie sich bitte an den Hersteller! FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Abbildung.13 FALSCH DE-7 Abbildung.14 KORREKT 3.1.4 Beladen und überladen der Rotoren Alle zugelassenen Rotoren werden mit ihrer Höchstgeschwindigkeit und dem maximalen Füllgewicht in "Tabelle 2 Zulässiges Nettogewicht" (siehe ANHANG) aufgeführt. Die für einen Rotor zulässige Höchstbelastung, die vom Hersteller bestimmt wird, sowie die Höchstgeschwindigkeit für diesen Rotor (siehe Etikett am Rotor) darf nicht überschritten werden. Die Flüssigkeiten, mit denen die Rotoren beladen werden, sollten eine maximale homogene Dichte von 1,2 g/ml oder weniger haben, wenn der Rotor mit maximaler Geschwindigkeit läuft. Um Flüssigkeiten mit einer höheren Dichte zu drehen, muss die Geschwindigkeit nach folgender Formel reduziert werden: Reduzierte Geschwindigkeit nred = 1,2 √ℎ𝑖𝑔ℎ𝑒𝑟 𝑑𝑒𝑛𝑠𝑖𝑡𝑦 x max. Geschwindigkeit (nmax) des Rotors Beispiel: nred = √ 1,2 x 4.000 = 3.360 rpm 1,7 Bei Fragen wenden Sie sich bitte an den Hersteller! Um die relative Zentrifugalkraft (RCF/G-Kraft) für einen bestimmten Adapter zu bestimmen, können Sie sie nach DIN 58 970 mit der beigefügten Formel berechnen: -5 2 RCF = 1.117862*10 *n *rmax n: Umdrehungen pro Minute (RPM) rmax: max. Zentrifugationsradius in cm bei Verwendung der Röhrenböden 3.1.5 Entfernung des Rotors Lösen Sie die Rotor-Befestigungsmutter vollständig (über den Steifpunkt schrauben) und heben Sie den Rotor vertikal aus der Zentrifuge heraus. (Siehe Abbildung 9 und 10). 3.2 Kontrolle des Deckels 3.2.1 Deckel öffnen Nach dem Lauf erscheint, wenn der Deckel der Zentrifuge offen ist, das Wort "schließen" (M1) auf der Anzeige "RPM | RCF" (A-1). Zusätzlich, wenn sich ein Rotor in der Zentrifuge befindet, erscheint das Wort "Rotor"(M3) wie auch die Code-Nummer des entsprechenden Rotors, der sich im Zentrifugensystem befindet "71" (M4). Wenn sich kein Rotor in der Zentrifuge befindet, blinkt das Wort "Rotor" (M3) und es erscheint zusätzlich das Wort"kein" (M4). Durch Drücken der Taste "Tür öffnen" (7) können Sie den Deckel der Zentrifuge freigeben. Sobald der elektromagnetische Deckel vollkommen freigegeben ist, erscheint das Wort "offen" (M2). Nun können Sie den Deckel der Zentrifuge öffnen. Bitte siehe Abbildung 15 unten zwecks Referenz DE-8 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R M2 M4 M3 M1 A-1 1 7 Abbildung. 15 Während des Betriebs können Sie den Rotortyp zu jeder Zeit durch Drücken der Taste "Tür öffnen" (7) aufrufen. 3.2.2 Deckelverschluss Der Deckel sollte nur leicht herabgedrückt werden. Ein elektromagnetischer Deckelverschluss schließt den Deckel, gleichzeitig verschwindet das Wort "offen" (M2) (siehe Abbildung 15). Als ein Zeichen dafür, dass die Zentrifuge Startbereit ist, erscheint in der Anzeige "RPM | RCF" (A-1) das Wort "schließen" (M1). Gleichzeitig wird auch das Wort "Rotor" (M3) angezeigt, wie auch die Codenummer des Rotors, die sich im Zentrifugensystem, "Nr 71" (M4) befindet. Damit werden alle rotorspezifischen Daten, wie z.B. maximale Geschwindigkeit, Beschleunigung usw. übernommen ACHTUNG: Greifen Sie mit Ihren Fingern nicht zwischen den Deckel und das Gerät oder den Sperrmechanismus, wenn Sie den Deckel verschließen! 3.3 Vorauswahl 3.3.1 Vorauswahl der Geschwindigkeit / RCF-Wert Diese Vorauswahl wird durch die Taste "RPM | RCF" (4) aktiviert (siehe Abbildung 16 unten). Wenn Sie die Taste einmal drücken, blinkt das Wort "rpm" (M5). Wenn Sie die Taste zweimal drücken, kann die Vorauswahl der Zentrifugalkräfte ausgewählt werden. Dann erscheint das blinkende Wort "rcf" (M6). Sie können die gewünschten Werte mit dem Anpassungsknopf (1) einstellen. In der Anzeige (A-1) wird stets der regulierte Wert angezeigt, bevor, M5 während und nach dem Betrieb. M6 A-1 1 4 5 Abbildung 16 Solange kein Rotor eingesetzt ist, kann die Geschwindigkeit zwischen 200 rpm und der maximalen Umdrehungszahl der Zentrifuge eingestellt werden. Wenn ein Rotor in der Zentrifuge vorhanden ist, kann die Geschwindigkeit nur bis zur maximal zulässigen Umdrehungszahl dieses Rotors vorgewählt werden. Genauso verhält es sich mit der Vorauswahl des RCF-Wertes. Der Einstellbereich liegt zwischen der Fliehkraft bei 200 U / min und der maximal zulässigen Fliehkraft des Rotors. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-9 Siehe "Tabelle 3: Max. Geschwindigkeit und RCF-Werte für zulässigen Rotor" (Siehe ANHANG). Dort sind alle wichtigen Werte aufgeführt. ACHTUNG: Bitte prüfen Sie auch die maximal zulässigen Drehzahlen Ihrer Teströhrchen mit dem Hersteller. 3.3.2 Vorauswahl der Laufzeit Die Laufzeit kann in drei verschiedenen Reichweiten von 10 Sekunden bis zu 99 Stunden 59 Minuten vorausgewählt werden. 1. Reichweite von 10 Sekunden bis zu 59 Minuten 50 Sekunden in Schritten von 10 Sekunden 2. Reichweite von 1 Stunde bis zu 99 Stunden 59 Minuten in Schritten von 1 Minute. 3. Der dauerhafte Betrieb "cont", der durch die Taste "Stop"(10) unterbrochen werden kann (siehe Abbildung 17). Die Betriebszeit kann vorab ausgewählt werden, während der Deckel offen oder geschlossen ist. Um die Einstellung der Betriebszeit einzustellen, drücken Sie die Taste "time" (6). Auf der Anzeige "Time/Prog" (A-3) blinkt die Anzeige "m : s" oder "h : m", je nach vorheriger Einstellung. Um den gewünschten Wert einzustellen, benutzen Sie den Anpassknopf (1). Nachdem 59 min 50 sec überschritten werden, wechselt die Anzeige automatisch auf "h : m". Nachzem 99 hours 59 min überschritten werden, erscheint das Wort "cont" in der Anzeige "Time/Prog" (A-3). Dieser dauerhafte Betrieb kann nur durch Drücken der Taste "Stop" (10) unterbrochen werden. Der Countdown beginnt so bald die eingestellte Geschwindigkeit erreicht ist. Die Anzeige zeigt immer die verbliebene Zeit. (Siehe Abbildung 17) A-3 1 10 6 Abbildung 17 3.3.3 Vorauswahl der Bremsintensität und Beschleunigung Diese Funktion wird durch die Taste "Accel/Decel" (5) aktiviert (siehe Abbildung18). Wenn die Taste nur einmal gedrückt wird, blinkt das Wort "accel" (M7) in der Anzeige "Acc/Dec" (A-2). Die gewünschte Beschleunigung kann vorab durch den Anpassungsknopf (1) ausgewählt werden. Der Wert 0 entspricht der geringsten und der Wert 9 der höchsten Beschleunigung. Durch zweimaliges Drücken der Taste "Accel/Decel" (5) zeigt die Anzeige "Acc/Dec" (A-2) das Wort "decel"(M8) an. Nun kann die gewünschte Bremsintensität vorab durch den Anpassungsknopf (1) ausgewaehlt werden. Der Wert 9 entspricht der kürzesten und der Wert 0 der längsten möglichen Bremszeit. Siehe "Tabelle 5: Beschleunigungs- und Verlangsamungszeiten" (ANHANG). Dort werden die Beschleunigungs- und Verlangsamungszeiten für die Beschleunigungs- und Verzögerungsphasen 0-9 für zulässige Rotoren gezeigt. M7 M8 A-2 1 5 Abbildung. 18 DE-10 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 3.3.4.Vorauswahl der Temperatur (Nur gekühlte Modelle) Diese Funktion wird von der Taste "Temp/Setup" (13) aktiviert. Nach dem Drücken dieser Taste blinkt in der Anzeige "Time/Prog" die Anzeige "°C" (A-4). (1) die gewünschte Testtemperatur kann in Schritten von 1°C in einer Reichweite von -20°C bis zu +40°C vorab ausgewählt werden. Der Wert wird ständig auf der Anzeige angegeben (Abbildung 19) - vor, während und nach dem Betrieb. Bitte beachten Sie die entsprechenden, niedrigsten Temperaturen der Rotoren bei maximaler Geschwindigkeit! A-4 1 13 Abbildung 19 3.3.5 Vorkühlung (nur gekühlte Modelle) Wenn die Proben temparaturempfindlich sind, ist es sinnvoll, die Zentrifuge, den Rotor und möglicherweise die Behälter auf die gewünschte Arbeitstemperatur vorzukühlen. Fügen Sie daher den gewünschten Rotor ein und stellen Sie die entsprechende Temperatur vorab ein. Durch gleichzeitiges Drücken der Tasten "Temp/Setup" (13) (siehe Abbildung 20) und "Time" (6) starten Sie den Durchlauf. Während des Betriebs waehlt die Einheit automatisch eine Rotationsgeschwindigkeit, die 20% der zulässigen Rotationsgeschwindigkeit des entsprechenden Rotors entspricht. Nachdem die Vorstarttemperatur erreicht ist, können Sie den Vorkühldurchlauf mit der "Stop"Taste (10) verlassen. Je nach eingefügtem Rotor, dauert die Vorkühlung ungefähr zwischen 10 und 20 min. A-4 13 10 6 Abbildung. 20 3.4 Radiuskorrektur Wenn Sie Adapter oder Reduzierer benutzen, könnte es den Zentrifugalradius des entsprechenden Rotors verändern. In diesem Fall können Sie den korrekten Radius manuell korrigieren. Gehen Sie bitte wie folgt vor: Schliessen Sie den Deckel, drücken Sie dann die Taste "Time" (6) (siehe Abbildung 21) und die Taste "Prog/Setup" (11) zur gleichen Zeit und halten Sie sie. Auf der Anzeige "Time/Prog" (A-3) erscheint das Wort "radius" (M9). Durch den Anpassungsknopf (1) können Sie die entsprechende Radiuskorrektur in Schritten von 0,1 cm (Siehe Tabelle 7, ANHANG) vorab auswählen. Sobald Sie die Radiuskorrektur eingestellt haben, erscheint das Wort "Radius" (M9). Das Wort wird sichtbar sein, bis Sie die Radiuskorrektur auf 0 zurücksetzen. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-11 A-3 M9 11 1 6 Abbildung. 21. 3.5 Programm 3.5.1 Speicherung von Programmen Sie können bis zu 99 Durchläufe mit allen relevanten Parametern speichern, einschließlich der benutzten Rotoren. Sie können jede freie Programmnummer benutzen, und sie erneut aufrufen. Setzen Sie den erforderlichen Rotor in die Zentrifuge. Durch Drücken der Taste "Prog/Setup" (11) erscheint auf der Anzeige "Time/Prog" (A-3) das Wort "programm". Mit dem Anpassungsknopf (1) können Sie die gewünschte Programmnummer auswählen. Wenn eine Programmnummer bereits belegt ist, erscheinen auf der Anzeige "RPM | RCF" (A-1) die Worte "Rotor" (M3) und "xx" (M4). Im Falle einer freien Programmnummer erscheint 0 M4 A-1 M3 A-3 M10 11 1 12 Abbildung. 22 Schließen Sie den Deckel der Zentrifuge. Fahren Sie nun fort wie zuvor beschrieben, um alle wichtigen Durchlaufparameter einzustellen. Wenn der Deckel nicht geschlossen ist, wenn Sie das Programm speichern, blinken auf der Anzeige "RPM/RCF" (A-1) die Worte "FirSt" und "CLOSE Lid" (siehe Abbildung 23) abwechselnd. Wenn Sie den Durchlauf starten möchten, ohne das Programm zu speichern, blinken auf der Anzeige "RPM/RCF" (A-1) die Worte "First" und "PrESS StoreE" (siehe Abbildung 24) abwechselnd. Abbildung 23 Abbildung. 24 DE-12 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Zur Anpassung der Daten drücken Sie die Taste "Store" (12) (siehe Abbildungen 23 und 24) für ungefähr 1 Sekunde. Wenn das Programm korrekt gespeichert wird, erscheint das Wort "StorE" auf der Anzeige "RPM/RCF" (A-1). Als ein Ergebnis verschwindet das Wort "programm" (M10). Sobald die Taste "Store" (12) freigegeben wird, erscheint das Wort "programm xx" (M10) erneut – das (xx) steht für den ausgewählten Programmstandort. Wenn die Programmnummern belegt sind, können Sie eine alte Nummer nehmen, die nicht mehr benötigt wird und nur die neuen Parameter eingeben. 3.5.2 Aufruf von gespeicherten Programmen Um gespeicherte Programme aufzurufen, drücken Sie die Taste "Prog/Setup" (11) (siehe Abbildung 25), während der Deckel bereits geschlossen ist. Innerhalb der Anzeige "Time/Prog" (A-3) erscheint "programm –"(M10). Die gewünschte Programmnummer kann vorab mit dem Anpassungsknopf (1) ausgewählt werden. Auf den entsprechenden Anzeigen erscheinen die gespeicherten Werte für dieses Programm. Wenn sich der falsche Rotor für das vorab ausgewählte Programm innerhalb der Zentrifuge befindet blinkt auf der Anzeige "RPM | RCF" (A-1) das Wort "rotor" (M3). Gleichzeitig werden das Wort "FALSE" und die gespeicherte Rotornummer " xx" (M4) abwechselnd blinken. M3 M4 A-3 A-1 M10 11 1 12 Abbildung. 25 3.5.3 Verlassen des Programmmodus Um den Programmmodus zu verlassen, drücken Sie lediglich die Taste "Prog/Setup" (11) (siehe Abbildung 25). Dann erscheint innerhalb der Anzeige "Time/Prog" das Wort "programm" . Stellen Sie die Anzeige auf "programm--" (M10), indem Sie den Anpassungsknopf benutzen (1). 3.6 Starten und Stoppen der Zentrifuge 3.6.1 Starten der Zentrifuge Sie können die Zentrifuge entweder mit der Taste"Start" (9) (siehe Abbildung 26) oder der Taste "Quick Spin" (8) starten. Mit der Taste "Start" (9) können Sie gespeicherte Durchläufe oder Durchläufe mit manuell vorab ausgewählten Parametern starten. Wenn die entsprechende, vorab ausgewählte Laufzeit abgelaufen ist, wird die Zentrifuge automatisch anhalten. Mit der Taste "Quick Spin" (8) können Sie Durchläufe starten, die nur einige Sekunden dauern werden. Durch Drüken der Taste "Quick Spin" (8) beschleunigt die Zentrifuge auf die vorab ausgewählte Umdrehungszahl. Auf der Anzeige "Time/Prog" (A-3) wird die abgelaufene Zeit ab dem Datum angezeigt, an dem die Taste "Quick Spin" gedrückt wurde (8). Durch Freigabe der Taste "Quick Spin" (8) stoppt die Zentrifuge und die Laufzeit wird bis zum Öffnen des Deckels angezeigt. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-13 A-3 9 1 8 Abbildung. 26 3.6.2 Die "STOP"-Taste Mit der Taste "Stop" (10) (siehe Abbildung 27) können Sie den Durchlauf jederzeit unterbrechen. Nachdem Sie die Taste gedrückt haben, wird sich die Zentrifuge mit der vorab ausgewählten Intensität bis zum Stillstand verlangsamen. 10 Abbildung. 27 3.7 Feststellung eines Ungleichgewichts Für den Fall, dass der Rotor nicht gleichmäßig beladen ist, wird der Antrieb sich bei der Beschleunigung ausschalten. Der Rotor verlangsamt sich bis zum Stillstand. Wenn in der Anzeige "Time/Prog" (A-3) das Wort "error” (M11) zusammen mit der Zahl "01” erscheint, ist der Gewichtsunterschied zwischen den Proben zu groß. Verteilen Sie das Gewicht gleichmäßig. Beladen Sie den Rotor wie in Kapitel 3.1.2 und 3.1.3 beschrieben. Wenn auf der Anzeige "Time/Prog" (A-3) das Wort "error” zusammen mit der Zahl "02" (siehe Abbildung 28) erscheint, könnte dies folgenden Grund haben: Der Schalter für Ungleichgewicht ist defekt. A-3 M11 Abbildung 28 DE-14 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4. EINSTELLUNG 4.1 Grundanpassungen 4.1.1 Zugriff auf den Modus "Betriebsdaten" Wenn Sie die Zentrifuge benutzen, können die folgenden Parameter eingestellt werden: - Temperaturanzeige °C oder °F - Akustisches Signal ein-/ausschalten - Tastaturgeräusch ein-/ ausschalten - Lautstärke-Vorauswahl des Klangsignals - Melodieauswahl des Klangsignals "Ende des Durchlaufs" Waehrend die Zentrifuge ausgeschaltet ist, drücken Sie gleichzeitig die Tasten "Time"(6) und "Door Open" (7) und schalten Sie die Zentrifuge am Hauptschalter ein. Lassen Sie nun beide Tasten los und als ein Ergebnis hiervon wird ein Anzeigetest von ungefähr 5 Sekunden durchgeführt. Alle Indikatoren werden gleichzeitig erscheinen (siehe Abbildung 29). 7 6 Abbildung. 29 ACHTUNG: Bitte beachten Sie, dass Sie das Programm wie unter Punkt 4.1.1 beschrieben eingeben müssen, um die Anpassungen der Punkte 4.1.2 - 4.1.7 zu ändern. Nachdem Sie die Einstellungen gespeichert haben, können Sie den normalen Programmiermodus wieder ändern, indem Sie die Zentrifuge kurzzeitig ausschalten. Alle geänderten Einstellungen müssen mit der Taste "Start"(9) bestätigt werden. Das Wort "Store"(12) erscheint auf der Anzeige "RPM | RCF"(A-1) - Nur dann sind die Vorauswahlen gültig! A-1 9 12 Abbildung. 30 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-15 4.1.2 Temperaturangabe Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann die Taste "Accel/Decel" (5). Auf der Anzeige "Acc/Dec" (A-2) erscheint das Wort "Service". Waehlen Sie nun den Buchstaben "C" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen auf der Anzeige "RPM | RCF" (A-1) die Worte "CELSI/temp" . Wenn Sie die Taste "RPM | RCF" (4) drücken, blinkt das Wort "CELSI" und Sie können mit dem Anpassungsknopf (1) die Anzeige in Fahrenheit "FAREN" ändern (siehe Abbildung 31). Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), ändern Sie zurück zum normalen Programmiermodus, indem Sie die Zentrifuge kurzzeitig ausschalten. 1 4 5 Abbildung. 31 4.1.3 Signal ein-/ausschalten Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann die Taste "Accel/Decel"(5). Auf der Anzeige blinkt "Acc/Dec" (A-2) das Wort "Service" . Wählen Sie nun den Buchstaben "L" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen die Worte "On Sound" auf der Anzeige "RPM | RCF" (4). Wenn Sie die Taste "RPM | RCF" (4) nun drücken, blinkt das Wort "On" und Sie können den Ton mit dem Anpassungsknopf (1) umschalten (siehe Abbildung 32). Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), ändern Sie zurück zum normalen Programmiermodus, indem Sie die Zentrifuge kurzzeitig ausschalten. A-1 A-2 1 4 5 Abbildung. 32 DE-16 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4.1.4 Lautstärke-Vorauswahl des Tonsignals Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann die Taste "Accel/Decel" (5). Auf der Anzeige "Acc/Dec" (A-2) blinkt das Wort "Service" . Wählen Sie nun den Buchstaben "U" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen auf der Anzeige "RPM | RCF" (A-1) die Worte "Vol=0- 9/Sound" . Nach dem Drücken der Taste "RPM | RCF" (4), können Sie die gewünschte Lautstaerke zwischen 0 (leise) und 9 (laut) mit dem Anpassungsknopf (1) anpassen (siehe Abbildung 33). Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), können Sie zurückwechseln zum normalen Programmiermodus, indem Sie die Zentrifuge für einen kurzen Zeitraum ausschalten. A-1 A-2 1 4 5 Abbildung. 3 4.1.5 Liedauswahl für das Tonsignal am Ende des Durchlaufs Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann die Taste "Accel/Decel" (5). Auf der Anzeige blinkt "Acc/Dec" (A-2) das Wor t"Service" . Wählen Sie nun den Buchstaben "G" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen auf der Anzeige "RPM | RCF" (A-1) die Worte "SonGo/Sound" . Nach dem Drücken der Taste "RPM | RCF" (4) können Sie mit Hilfe des Anpassungsknopfes (1) ein Lied auswählen. (Siehe Abbildung 34). Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), können Sie zurückwechseln zum normalen Programmiermodus, indem Sie die Zentrifuge kurzzeitig ausschalten. A-1 A-2 1 4 5 Abbildung. 34 4.1.6 Tastaturgeräusch ein-/ ausschalten Fahren Sie fort wie unter Punkt 4.1.1 beschrieben, um in den Programmmodus einzutreten und drücken Sie dann die Taste "Accel/Decel" (5). Auf der Anzeige "Acc/Dec" (A-2) blinkt das Wort "Service" . Wählen Sie nun den Buchstaben "B" mit Hilfe des Anpassungsknopfes (1). Als Ergebnis erscheinen auf der Anzeige "RPM | RCF" (A-1) die Worte "ON/BEEP" . Nach dem Drücken der Taste "RPM | RCF" (4) können Sie das Tastaturgeräusch mit dem Anpassungsknopf (1) ein- (On) oder ausschalten (Off). (Siehe Abbildung 39). Nachdem Sie die Einstellungen gespeichert haben (siehe 4.1.1), können Sie zurückwechseln zum normalen Programmiermodus, indem Sie die Zentrifuge kurzzeitig ausschalten. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-17 A-2 A-1 1 4 5 Abbildung. 35 4.1.7 Betriebsdaten aufrufen ACHTUNG: Dies sollte nur durch einen fortgeschrittenen Benutzer oder Servicetechniker ausgeführt werden. Im Modus "Grundanpassungen" können Sie die Betriebsdaten der Zentrifuge aufrufen. Fahren Sie bitte wie unter Punkt 4.1.2 beschrieben fort, um in diesen Programmmodus einzutreten. Drücken Sie die Taste "Accel/Decel" (5). Auf der Anzeige "Acc/Dec" (A-2) blinkt das Wort "Service" . Mit dem Anpassungsknopf (1) können die verschiedenen Informationen erreicht werden: A= vorherige Starts der Zentrifuge H= vorherige Betriebsstunden S= Softwareversion r= Konvertersoftware E= Liste der vorherigen Fehlermitteilungen h= Laufzeit des Motors Die Liste der letzten 99 Fehlermitteilugen kann aufgerufen werden, indem die Taste "RPM | RCF" (4) gedrückt wird, und Sie können mit Hilfe des Anpassungsknopfes (1) durchscrollen. Die entsprechenden Fehlercodes erscheinen in der Anzeige "RPM | RCF" (A-1). Siehe bitte "Tabelle 6: Fehlermitteilungen" (siehe ANHANG). Um auf den normalen Programmmodus zu wechseln, schalten Sie die Zentrifuge für eine kurze Zeit aus. A-2 A-1 1 4 5 Abbildung 36 DE-18 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 5. WARTUNG 5.1 Wartung und Reinigung 5.1.1 Allgemein Pflege: Die Wartung der Zentrifuge ist darauf beschränkt, den Rotor, die Rotorkammer und die Rotorzubehörteile sauber zu halten sowie die Rotoreinsatzschrauben eines Ausschwenkrotors (falls vorhanden) regelmäßig zu schmieren. Das geeignetste Schmiermittel ist das High TEF-Öl. Schmiermittel, die Molycote und Graphit enthalten, sind nicht zugelassen. Achten Sie vor allem auf die eloxierten Aluminiumteile. Sogar der leichteste Schaden kann zu einem Bruch der Rotoren führen. Für den Fall dass Rotoren, Behälter oder Rohrgestelle mit korrosiven Substanzen in Kontakt stehen, müssen die betroffenen Stellen sorgfältig gereinigt werden. Korrosive Substanzen sind zum Beispiel: Alkalien, Basen, Seifenlösungen, Alkaliamine, konzentrierte Säuren, Lösungen, die Schwermetalle enthalten, wasserfreie, chlorhaltige Lösungsmittel, Salzlösungen, z.B. Salzwasser, Phenol, halogenierte Kohlenwasserstoffe. Reinigung - Einheiten, Rotoren, Zubehörteile - Schalten Sie das Gerät aus und trennen Sie es von der Stromversorgung, bevor Sie mit irgendwelchen Reinigungs- und Desinfektionsarbeiten beginnen. Schütten Sie keine Flüssigkeiten in das Innere des Gehäuses. Sprühen Sie kein Desinfektionsmittel auf das Gerät. Eine gründliche Reinigung erfüllt nicht nur seinen Zweck hinsichtlich der Hygiene, sondern vermeidet auch Korrosion aufgrund von Verschmutzung. Um eine Beschädigung von eloxierten Teilen wie z.B. Rotoren, Reduktionsplatten, usw. zu vermeiden, benutzen Sie nur pH-neutrale Reinigungsmittel mit einem pH-Wert von 6-8 zur Reinigung. Basische Reinigungsmittel (pH-Wert > 8) dürfen nicht benutzt werden. Stellen Sie nach der Reinigung sicher, dass alle Teile gründlich getrocknet sind, entweder per Hand oder in einer Heißluftkammer (max. Temperatur + 50°C). Es ist erforderlich, die eloxierten Aluminiumteile regelmäßig mit einem Korrosionsschutz-Öl zu beziehen, um ihre Lebensdauer zu erhöhen und die Korrosionsanfälligkeit zu reduzieren. Aufgrund von Feuchtigkeit oder nicht hermetisch verriegelten Proben können sich Kondensate bilden. Das Kondensat muss von der Rotorkammer regelmäßig mit einem weichen Tuch entfernt werden. Das Wartungsverfahren muss alle 10 bis 15 Durchläufe oder mindestens einmal pro Woche wiederholt werden. - Schließen Sie die Einheit an die Stromversorgung an, nachdem die Ausrüstung vollkommen trocken ist. Führen Sie keine Desinfektion mit UV-, Beta- und Gammastrahlen oder anderer hochenergetischer Strahlung durch. Metallrotoren können autoklaviert werden. Rotordeckel und Adapter können ebenfalls autoklaviert werden (max. 121°C, 20 min). Die Reagenzglasgestelle sind aus PP und können nicht bei 134°C autoklaviert werden. 5.1.2 Reinigung und Desinfektion der Einheit 1. Öffnen Sie den Deckel bevor Sie die Einheit ausschalten. Trennen Sie sie von der Stromversorgung. 2. Öffnen Sie die Rotormutter, indem Sie den Rotorschlüssel im Uhrzeigersinn drehen. 3. Entfernen Sie den Rotor. 4. Benutzen Sie zur Reinigung und Desinfektion der Einheit und der Rotorkammer das oben genannte Reinigungsmittel. 5. Reinigen Sie alle zugänglichen Bereiche des Geräts und seine Zubehörteile, einschließlich des Stromkabel mit einem feuchten Tuch. 6. Waschen Sie die Gummidichtungen und die Rotorkammer gründlich mit Wasser. 7. Schrubben Sie die Gummidichtungen mit Glyzerol oder Talkum trocken, um zu vermeiden, dass sie spröde werden. Andere Komponenten der Einheit, z.B. der Deckelverschluss, die Motorwelle und der Rotor dürfen nicht eingefettet werden. 8. Trocknen Sie die Motorwelle mit einem weichen, trockenen und fuselfreien Tuch. 9. Kontrollieren Sie die Einheit und die Zubehörteile auf Beschädigungen. Stellen Sie sicher, dass die Zentrifuge abgeschaltet und trennen Sie die Einheit von der Stromversorgung. Entfernen Sie dann den anhaftenden Staub von den Ventilationsschlitzen in der Zentrifuge, indem Sie eine weiche Bürste verwenden. Wiederholen Sie dies mindestens alle 6 Monate. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-19 5.1.3 Reinigung und Desinfektion des Rotors 1. Reinigen und desinfizieren Sie die Rotoren, Rotordeckel und Adapter mit dem oben genannten Reinigungsmittel. 2. Benutzen Sie eine Flaschenbürste, um die Rotorbohrungen zu reinigen und zu desinfizieren. 3. Spülen Sie die Rotoren, den Rotordeckel und den Adapter mit klarem Wasser ab. Vor allem die Bohrungen der Winkelrotoren. 4. Um die Rotoren und Zubehörteile zu trocknen, legen Sie sie auf ein Handtuch. Stellen Sie die Winkelrotoren mit den Bohrungen nach unten auf das Handtuch. 5. Trocknen Sie den Kegel mit einem weichen, trockenen und fuselfreien Tuch und halten Sie Ausschau nach Beschädigungen. Fetten Sie den Rotorkegel nicht ein. 5.1.4 Desinfektion von Aluminiumrotoren Im Falle von infektiösen Materialien, die in die Zentrifuge auslaufen, müssen der Rotor und die Rotorkammer direkt nach dem Durchlauf desinfiziert werden. Rotoren können bei einer Maximaltemperatur von 121°C autoklaviert werden. 5.1.5 Desinfektion der PP-Rotoren Autoklavieren Die empfohlene Zeit fürs Autoklavieren: 15 – 20 min bei 121°C (1 bar) ACHTUNG: Die Sterilisationszeit von 20 min. darf nicht überschritten werden. Eine wiederholte Sterilisation wird zu einer Verminderung der mechanischen Widerstands des Kunststoffmaterials führen Vor dem Autoklavieren müssen der PP-Rotor und der Adapter gründlich gereinigt werden, um ein Verbrennen von Schmutzresten zu vermeiden. Sie können die Auswirkungen einiger chemikalischen Reste auf Kunststoffmaterialien bei Umgebungstemperaturen ignorieren. Aber bei den hohen Temperaturen beim Autoklavieren können diese Reste korrodieren und den Kunststoff zerstören. Die Objekte müssen nach der Reinigung aber vor dem Autoklavieren gründlich mit destilliertem Wasser abgespült werden. Reste von Reinigungsflüssigkeiten können Spalten, Aufhellungen und Flecken hervorrufen. Gassterilisation Adapter, Flaschen und Rotoren kann mit Ethylenoxid gassterilisiert werden. Stellen Sie sichern, die Objekte nach der Sterilisation und vor der erneuten Verwendung lüften. ACHTUNG: Da die Temperatur bei der Sterilisation ansteigen kann, dürfen Rotoren, Adapter und Flaschen nicht geschlossen werden, und müssen vollkommen aufgeschraubt sein. Chemische Sterilisation Flaschen, Adapter und Rotoren können mit den üblichen flüssigen Desinfektionsmitteln benutzt werden. ACHTUNG: Bevor Sie eine andere Reinigungs- oder Dekontaminationsmethode als vom Hersteller empfohlen anwenden, nehmen Sie mit dem Hersteller Kontakt auf, um sicherzustellen, dass dies die Einheit oder den Rotor nicht beschädigen wird. 5.1.6 Glasbruch Bei hohen G-Werten erhöht sich die Bruchrate von Reagenzgläsern aus Glas. Glassplitter müssen unverzüglich vom Rotor, den Behälter und Adaptern sowie der Rotorkammer selbst entfernt werden. Feine Glassplitter werden die Schutzbeschichtung eines Rotors kratzen und daher beschädigen. Wenn Glassplitter auf der Rotorkammer verbleiben, wird sich aufgrund der Luftzirkulation Feinmetallstaub entstehen. Dieser sehr feine, schwarze Metallstaub wird die Rotorkammer, den Rotor, die Behälter und die Proben erheblich verschmutzen. Tauschen Sie, wenn erforderlich, Adapter, Rohre und Zubehörteile aus, um weitere Schäden zu vermeiden. Prüfen Sie die Rotorlöcher regelmäßig auf Ablagerungen und Beschädigungen. ACHTUNG: Bitte prüfen sie die entsprechenden Spezifikation der Reagenzglaszentrifugen mit dem Hersteller. 5.2 Lebenszeit der Rotoren, Behälter, Zubehörteile Rotoren und Rotordeckel aus Aluminium oder rostfreiem Stahl verfügen über eine Betriebszeit von max. 7 Jahren ab dem ersten Gebrauch. Transparente Rotordeckel und Kappen aus PC oder PP sowie Rotoren, Reagenzglasgestelle und Adapter aus PP verfügen über eine maximale Betriebsdauer von bis zu 3 Jahren ab dem ersten Gebrauch. Bedingungen für die Betriebszeit: Ordnungsgemäße Anwendung, schadenfreier Zustand, empfohlene Pflege DE-20 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 6. PROBLEMBEHEBUNG 6.1 Fehlermitteilung: Ursache / Lösung Die Fehlermitteilungen werden aufgelistet, um Ihnen dabei zu helfen, mögliche Fehler schneller zu finden. Die Diagnose, auf die in diesem Kapitel Bezug genommen wird, kann nicht immer zutreffen, da es sich nur um theoretisch auftretende Fehler und Lösungen handelt. Bitte halten Sie uns über alle möglichen auftretenden Fehler auf dem Laufenden, der in diesem Kapitel nicht aufgeführt wird. Nur durch Ihre Informationen sind wir in der Lage, dieses Betriebshandbuch zu verbessern. Vielen Dank im Voraus für Ihre Unterstützung. 6.2 Untersuchung der möglichen Fehlermitteilungen und ihrer Lösungen 6.2.1 Deckelfreigabe bei Stromausfall (Notfalldeckelfreigabe) Im Falle eines Stromausfalls oder einer Fehlfunktion kann der Deckel der Zentrifuge manuell geöffnet werden, um Ihre Proben herauszunehmen. Für FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (motorgetriebenes Schloss) Gehen Sie bitte wie folgt vor: 1. Schalten Sie die Zentrifuge aus und ziehen Sie das Stromkabel heraus, warten Sie, bis der Rotor still steht (dies kann einige Minuten dauern) 2. Auf der linken Seite des Zentrifugengehäuses, gibt es einen Plastikstopper. Entfernen Sie diese Stopper, dahinter befindet sich eine sechseckige Mutter. 3. Nehmen Sie den Steckschlüssel, stecken Sie ihn in das Loch und schließen Sie den Steckschlüssel mit der sechseckigen Mutter (Siehe Abbildung 37). 4. Drehen Sie nun den Steckschlüssel bis zum Limit auf die rechte Seite (im Uhrzeigersinn). ACHTUNG: a) Drehen Sie nur bis zum Limit, ziehen Sie die Mutter nicht fest. b) Öffnen Sie nun den Deckel der Zentrifuge. c) Schalten Sie die Zentrifuge wieder ein, um die Arbeit wiederaufzunehmen. Abbildung. 37 Für 5714 Bitte gehen Sie wie folgt vor (siehe Abbildung 38): ACHTUNG: • Schalten Sie die Zentrifuge aus und ziehen Sie das Stromkabel heraus, warten Sie, bis der Rotor still steht (dies kann einige Minuten dauern) Auf der rechten Seite der Zentrifuge gibt es einen Plastikstopper (Abbildung 38). Entfernen sie diesen Stopper, der am Deckelverschluss angeschlossen ist, horizontal vom Gehäuse, bis sich der Zentrifugendeckel öffnet. • Öffnen Sie nun den Deckel der Zentrifuge FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-21 Abbildung. 38 6.2.2 Beschreibung des Fehlermitteilungssystems Die Fehlermitteilung "Error" (M11) wird auf der "Time/Prog" (A-3)-Anzeige angezeigt (siehe Abbildung 39). Detaillierte Informationen über mögliche Fehlermitteilungen sind enthalten in "Tabelle 6: Fehlermitteilungen" (Siehe Anhang). A-3 M11 Abbildung. 39 7. ENTGEGENNAHME VON ZU REPARIERENDEN ZENTRIFUGEN Gesundheitsrisiko von kontaminierter Ausrüstung, Rotoren und Zubehörteilen. Für den Fall, dass die Zentrifuge zur Reparatur an den Hersteller zurückgesandt wird, beachten Sie bitte Folgendes: Die Zentrifuge muss vor der Versendung dekontaminiert und gereinigt werden, um Personen, Umwelt und Material zu schützen. Dekontaminationszertifikat für die Rücksendung von Gütern (Siehe ANHANG) Wir behalten uns das Recht vor, kontaminierte Zentrifugen nicht anzunehmen. Zudem gehen alle Kosten, die bei der Reinigung und Desinfektion der Einheiten entstehen zu Lasten des Kunden. DE-22 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 8. TRANSPORT, LAGERUNG UND ENTSORGUNG 8.1 Transport Nehmen Sie den Rotor vor dem Transport heraus. Transportieren Sie die Einheit nur in der Originalverpackung. Installieren Sie das Transportschutzmaterial, um die Motorwelle zu schützen, wenn Sie das Gerät über längere Strecken transportieren. General transportation Air temperature rel. humidity Air pressure -25 to 60 °C 10 to 75 % 30 to 106 kPa 8.2 Lagerung Während der Lagerung der Zentrifuge müssen die folgenden Umweltbedingungen eingehalten werden: in transport packaging Air temperature rel. Humidity Air pressure -25 to 55 °C 10 to 75 % 70 to 106 kPa 8.3 Transport, Installation, Übertragung und Entsorgung der Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R Diese Anweisungen ergänzen die vorherigen allgemeinen Anweisungen in Kapitel 8 und ersetzen sie nicht 8.3.1 Transport • Bitte transportieren Sie das Gerät in der Originalverpackung. • Die Zentrifuge sollte immer unter Verwendung von mechanischem Transportgerät transportiert werden. 8.3.2 Installation  Öffnen des Kartons und Heben des Geräts 1. Schneiden Sie das Klebeband ab. 2. beschrieben.Öffnen Sie alle 4 Klappen des Kartons 3. Entfernen Sie die Zubehörteile 4. Heben Sie die Zentrifuge vorsichtig aus dem Karton. WARNUNG: Hebegefahr. Das Heben durch eine einzige Person kann Unfälle verursachen. Benutzen Sie ein mechanisches Hebegerät oder verknüpfen Sie Hebeverfahren, wenn Sie die Ausrüstung heben oder bewegen.  Platzieren Sie das Gerät auf einem stabilen, horizontalen und nicht vibrierenden Labortisch. 1. Entfernen Sie das vordere und hintere Schutzmaterial 2. Entfernen Sie die Plastikfolie. 3. Beachten Sie den Mindestabstand von 30 cm zu benachbarten Geräten an den Seiten und von der Rückseite bis zur Wand. 4. Installieren Sie das Gerät an einem gut gelüfteten Standort, der vor direktem Sonnenlicht geschützt ist, um eine Überhitzung zu vermeiden. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R /FC5916/FC5916R DE-23  Schließen Sie das Gerät an 1. Warten Sie nach der Installation ungefähr vier Stunden, bevor Sie die Zentrifuge einschalten, um so eine Beschädigung des Kompressors zu vermeiden. 2. Prüfen Sie, dass die Netzspannung und die Frequenz den Anforderungen auf der Namensplatte des Geräts (siehe Rückseite des Geräts) übereinstimmen, und schließen Sie das Gerät dann an die Stromversorgung an.  Entfernen Sie das Transportschutzmaterial von der Rotorkammer. 1. Schalten Sie das Gerät am Hauptstromschalter ein. 2. Öffnen Sie den Zentrifugendeckel, indem Sie die Öffnen-Taste drücken 3. Entfernen Sie das Transportschutzmaterial 4. Platzieren Sie den Rotor vertikal auf dem Motorschaft. 5. Drehen Sie die Rotormutter, indem Sie den Rotorschlüssel im Uhrzeigersinn drehen, bis die Rotormutter festgezogen ist.  Das Gerät ist nun gebrauchsfertig Bewahren Sie die Verpackung und jegliches Transportmaterial für die Versendung des Geräts zu einem späteren Zeitpunkt auf. 8.3.3 Verpackung Packen Sie die Zentrifuge in umgekehrter Reihenfolge. 8.3.4 Weitergabe des Geräts Wenn Sie die Ausrüstung an Dritte weitergeben, stellen Sie bitte sicher, dass Sie auch diese Betriebsanleitung beifügen. DE-24 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9. TECHNISCHE DATEN 9.1 Spezifikationen 9.1.1 Zentrifuge FC5714 Modell Geschwindigkeitsreichweite FC5714 200 rpm -14000 rpm;10 rpm/Satz Maximales RCF 18624 x g;10 x g/Satz Maximale Kapazität(Rotor) 4x100ml Temperaturreichweite (Nicht verfügbar) Laufzeit Luftkühlung 10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft Lärmpegel (je nach Rotor) Zulässige Dichte bei Maximalgeschwindigkeit Zulässige kinetische Energie Netzanschluss AC ≤ 63 ± 2 dB(A) 1,2 g/ml 5595 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % Aktueller Verbrauch 1,3 A 2,4 A Stromverbrauch 240 W 300 W Spannungsschwankung 362 x 493 x 330 mm Ausmaße (B × T × H) 14,3 x 19,4 x 13,0 in 30 kg Nettogewicht (ohne Rotor) 66 lb 580 x 490 x 460 mm Versandausmaße (B × T × H) 22,8 x 19,3 x 18,1 in 32,5 kg Versandgewicht (ohne Rotor) 72 lb Umgebungsbedingungen (EN/IEC 61010-1) Umwelt Höhe Umgebungstemperatur Max. relative Luftfeuchtigkeit Überspannungskategorie (IEC 60364-4443) Kontaminationsgrad nur zur Innenanwendung Bis zu einer Höhe von 2000m benutzbar 2°C bis 35 °C Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis zu 31°C, linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend bis 35°C. Schutzgrad Nicht geeignet für die Anwendung in gefährlichen Umgebungen. II 2 I FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-25 9.1.2 Zentrifuge FC5718 Modell FC5718 Geschwindigkeitsreichweite 200 rpm -18000 rpm;10 rpm/Satz Maximales RCF 23542 x g;10 x g/Satz Maximale Kapazität (Rotor) 4x100ml Temperaturreichweite (Nicht verfügbar) Laufzeit Lärmpegel Luftkühlung 10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft Lärmpegel (je nach Rotor) Zulässige Dichte bei Maximalgeschwindigkeit Zulässige kinetische Energie Netzanschluss AC ≤ 60 ± 2 dB(A) 1,2 g/ml 16672 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % Aktueller Verbrauch 2.0 A 4.0 A Stromverbrauch 455 W 475 W 408 x 499 x 351 mm 16,1 x 19,7 x 13,8 in 43 kg Spannungsschwankung Ausmaße (B × T × H) Nettogewicht (ohne Rotor) 95 lb 650 x 520 x 490 mm Versandausmaße (B × T × H) 25,6 x 20,5 x 19,3 in 53 kg Versandgewicht (ohne Rotor) 117 lb Umgebungsbedingungen (EN/IEC 61010-1) Umwelt Höhe Umgebungstemperatur Max. relative Luftfeuchtigkeit Überspannungskategorie (IEC 60364-4443) Kontaminationsgrad nur zur Innenanwendung Bis zu einer Höhe von 2000m benutzbar 2°C bis 35 °C Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis zu 31°C, linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend bis 35°C. Schutzgrad Nicht geeignet für die Anwendung in gefährlichen Umgebungen. II 2 I DE-26 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R /FC5916/FC5916R 9.1.3 Zentrifuge FC5718R Modell Geschwindigkeitsreichweite FC5718R 200 rpm -18000 rpm;10 rpm/Satz Maximales RCF 23542 x g;10 x g/Satz Maximale Kapazität(Rotor) 4x100ml Temperaturreichweite(Digital) -20~40℃ Laufzeit 10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft Lärmpegel (je nach Rotor) Zulässige Dichte bei Maximalgeschwindigkeit Zulässige kinetische Energie Netzanschluss AC ≤ 60 ± 2 dB(A) 1,2 g/ml 25111 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % Aktueller Verbrauch 3,0 A 6,0 A Stromverbrauch 660 W 660 W Spannungsschwankung Ausmaße (B × T × H) 407 x 731 x 359 mm 16,0 x 28,8 x 14,1 in 60 kg Nettogewicht (ohne Rotor) Versandausmaße (B × T × H) 132 lb 840 x 640 x 590 mm 33,1 x 25,2 x 23,2 in 77 kg Versandgewicht (ohne Rotor) 170 lb Umgebungsbedingungen (EN/IEC 61010-1) Umwelt Höhe Umgebungstemperatur Max. relative Luftfeuchtigkeit Überspannungskategorie (IEC 60364-4443) Kontaminationsgrad nur zur Innenanwendung Bis zu einer Höhe von 2000m benutzbar 2°C bis 35 °C Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis zu 31°C, linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend bis 35°C. Schutzgrad Nicht geeignet für die Anwendung in gefährlichen Umgebungen. II 2 I 9.1.4 Centrifuge FC5816 Modell Geschwindigkeitsreichweite FC5816 200 rpm -6000 rpm;50 rpm/Satz Maximales RCF 4427 x g;10 x g/Satz Maximale Kapazität(Rotor) 6 x 250ml Temperaturreichweite (Nicht verfügbar) Laufzeit Luftkühlung 10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft Lärmpegel (je nach Rotor) Zulässige Dichte bei Maximalgeschwindigkeit Zulässige kinetische Energie Netzanschluss AC ≤ 61 +2 dB(A) 1,2 g/ml 34363 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % Aktueller Verbrauch 2,4 A 4,2 A Stromverbrauch 530 W 520 W Spannungsschwankung Ausmaße (B × T × H) 446 x 538 x 354 mm 17,6 x 21,2 x 13,9 in 52 kg Nettogewicht (ohne Rotor) Versandausmaße (B × T × H) 115 lb 840 x 640 x 590 mm 33,1 x 25,2 x 23,2 in 77 kg Versandgewicht (ohne Rotor) 170 lb Umgebungsbedingungen (EN/IEC 61010-1) Umwelt Höhe Umgebungstemperatur Max. relative Luftfeuchtigkeit Überspannungskategorie (IEC 60364-4443) Kontaminationsgrad nur zur Innenanwendung Bis zu einer Höhe von 2000m benutzbar 2°C bis 35 °C Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis zu 31°C, linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend bis 35°C. Schutzgrad Nicht geeignet für die Anwendung in gefährlichen Umgebungen. II 2 I DE-28 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.5 Zentrifuge FC5816R Modell Geschwindigkeitsreichweite FC5816R 200 rpm -16000 rpm;10 rpm/Satz Maximales RCF 24325 x g;10 x g/Satz Maximale Kapazität(Rotor) 6 x 250ml Temperaturreichweite(Digital) -20~40℃ Laufzeit 10 sek. bis 99 h. 99 min. 59 sek. oder dauerhaft Lärmpegel (je nach Rotor) Zulässige Dichte bei Maximalgeschwindigkeit Zulässige kinetische Energie Netzanschluss AC ≤ 63 ± 2 dB(A) 1,2 g/ml 34363 Nm 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ± 10 % ± 10 % Aktueller Verbrauch 3,7 A 7,8 A Stromverbrauch 785 W 850 W Spannungsschwankung 723 x 538 x 354 mm Ausmaße (B × T × H) 28,5 x 21,2 x 13,9 in 77 kg Nettogewicht (ohne Rotor) 170 lb 840 x 640 x 590 mm Versandausmaße (B × T × H) 33,1 x 25,2 x 23,2 in 87 kg Versandgewicht (ohne Rotor) 192 lb Umgebungsbedingungen (EN/IEC 61010-1) Umwelt Höhe Umgebungstemperatur Max. relative Luftfeuchtigkeit Überspannungskategorie (IEC 60364-4443) Kontaminationsgrad nur zur Innenanwendung Bis zu einer Höhe von 2000m benutzbar 2°C bis 35 °C Max. relative Luftfeuchtigkeit 80 % für Temperaturen von bis zu 31°C, linear auf eine relative Luftfeuchtigkeit von 50 % abnehmend bis 35°C. Schutzgrad Nicht geeignet für die Anwendung in gefährlichen Umgebungen. II 2 I FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-29 9.1.6 Centrifuge FC5916 Modell Geschwindigkeitsbereich Maximale RCF Maximale Kapazität(Rotor) Temperaturbereich(N/A) FC5916 200 U/min -16000 U/min;10 U/min/Satz 24325 x g;10 x g/Satz 4x750ml Luftkühlung Laufzeit 10 sec bis 99 Std. 59 min 59 sec oder kontinuierlich Geräuschpegel (abhängig vom Rotor) Zulässige Dichte bei maximaler Geschwindigkeit Zulässige kinetische Energie ≤ 63 ± 2 dB(A) Umgebungstemperatur 2°C bis zu 35°C Max. relative Luftfeuchtigkeit Max. relative Luftfeuchtigkeit 80 % für Temperaturen bis 31°C, 1.2 g/ml 60629 Nm 230 V ~ 50/60 Netzanschluss AC 120 V ~ 50/60 Hz Hz Spannungsschwankung ± 10 % ± 10 % Aktueller Stromverbrauch 2.8 A 5.6 A Leistungsaufnahme 640 W 680 W 540 x 670 x 390 mm Maße (B × T × H) 21.3 x 26.4 x 15.4 in 85 kg Nettogewicht (ohne Rotor) 187 lb 780 x 670 x 590 mm Maße für den Versand (B × T × H) 30.7 x 26.4 x 23.3 in 98 kg Transportgewicht (ohne Rotor) 216 lb Umgebungsbedingungen (EN/IEC 61010-1) Bereich Nur für den Innenbereich Höhe Einsatz bis zu einer Höhe von 2000 m linear abnehmend auf 50 % relative Luftfeuchtigkeit bis 35°C. Überspannungskategorie (IEC 60364-4II 443) Grad der Verschmutzung 2 Schutzklasse I Nicht geeignet für den Einsatz in gefährlicher Bereiche. TM DE-30 Frontier Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.7 Centrifuge FC5916R Modell Geschwindigkeitsbereich Maximale RCF Maximale Kapazität(Rotor) FC5916R 200 U/min -16000 U/min;10 U/min/Satz 26331 x g;10 x g/Satz 4x750ml Temperaturbereich(Digital) -20~40℃ Laufzeit 10 sec bis 99 Std. 59 min 59 sec oder kontinuierlich Geräuschpegel (abhängig vom Rotor) Zulässige Dichte bei maximaler Geschwindigkeit Zulässige kinetische Energie Netzanschluss AC Spannungsschwankung Aktueller Stromverbrauch Leistungsaufnahme ≤ 63 ± 2 dB(A) Maße (B × T × H) Nettogewicht (ohne Rotor) Maße für den Versand (B × T × H) Transportgewicht (ohne Rotor) 1.2 g/ml 54458 Nm 230 V ~ 50/60 Hz ± 10 % 7.2 A 1630 W 730 x 670 x 390 mm 28.7 x 26.4 x 15.4 in 118 kg 260 lb 900 x 750 x 560 mm 40.0 x 29.5 x 22.0 in 137 kg 137 kg 120 V ~ 50/60 Hz ± 10 % 20 A 1750 W Umgebungsbedingungen (EN/IEC 61010-1) Bereich Nur für den Innenbereich Höhe Einsatz bis zu einer Höhe von 2000 m Umgebungstemperatur 2°C bis zu 35°C Max. relative Luftfeuchtigkeit Max. relative Luftfeuchtigkeit 80 % für Temperaturen bis 31°C, linear abnehmend auf 50 % relative Luftfeuchtigkeit bis 35°C. Überspannungskategorie (IEC 60364-4II 443) Grad der Verschmutzung 2 Schutzklasse I Nicht geeignet für den Einsatz in gefährlicher Bereiche. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-31 9.2 Zeichnungen und Ausmaße D H W D H W TM DE-32 Frontier Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Modell FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R B (mm / in.) 362 / 14.3 408 / 16.1 408 / 16.1 446 / 17.6 723 / 28.5 540 / 21.3 730 / 28.7 T (mm / in.) 493 / 19.4 499 / 19.7 731 / 28.8 538 / 21.2 538 / 21.2 670 / 26.4 670 / 26.4 H (mm / in.) 330 / 13.0 351 / 13.8 359 / 14.1 354 / 13.9 354 / 13.9 390 / 15.4 390 / 15.4 10. COMPLIANCE Die Erfüllung der folgenden Standards wird durch die entsprechende Kennzeichnung auf dem Produkt angegeben. Kennzeichn Standard ung Dieses Produkt entspricht den geltenden, harmonisierten Standards der EUDirektiven 2011/65/EU (RoHS), 2014/30/EU (EMC) und 2014/35/EU (LVD). The EU-Konformitätserklärung ist online verfüg bar unter www.ohaus.com/ce. Entsorgung In Übereinstimmung mit der Europäischen Direktive 2012/19/EU über W aste Electrical and Electronic Equipment (WEEE), darf dieses Gerät nicht mit dem häuslichen Abfall entsorgt werden. Dies gilt auch für Län der außerhalb der EU, in Bezug auf ihre spezifischen Anforderungen. Bitte entsorgen Sie dieses Produkt entsprechend lokalen Bestimmungen an Sammelpunkten, die für elektrische und elektronische Ausrüstungen spezifiziert sind. Wenn Sie Fragen haben, nehmen Sie bitte mit der verantwortlichen Behörde oder dem Händler Kontakt auf, von dem Sie dieses Gerät gekauft haben. Sollte dieses Gerät an andere Parteien weitergegeben werden (für den privaten oder beruflichen Gebrauch), muss auch auf den Inhalt dieser Bestimmung verwiesen werden. Für Anweisungen zur Entsorgung in Europa, siehe www.ohaus.com/weee. Vielen Dank für Ihren Beitrag zum Umweltschutz. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-33 11. ANHANG TABELLE 1: ZULÄ SSIGES NETTOGEWICHT TABELLE 2: NIEDRIGSTE TEMPERATUREN UND MAX.-GESCHWINDIGKEIT TABELLE 3: MAX. GESCHWINDIGKEIT UND RCF-WERTE FÜ R ZULÄ SSIGE ROTOREN TABELLE 4: BESCHLEUNIGUNGS- UND VERLANGSAMUNGSZEITEN TABELLE 5: FEHLERMITTEILUNGEN TABELLE 6 RADIUSKORREKTUR TABELLE 7: RÜ CKNAHMEFORMULAR / DEKONTAMINATIONSZERTIFIKAT TM DE-34 Frontier Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.1 Table 1: ZULÄSSIGES NETTOGEWICHT FC5916R FC5916 FC5816R FC5816 FC5718R Zulä ssiges Nettogewicht FC5718 Rotorbeschreibung FC5714 Rotor-Nummer Maximale Drehzahl ( x1000 U/min) 14.0 14.0 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 15.0 24x 3.4 g 15.0 16.0 16.0 16.0 30314820 Rotor, Sw ing out, 4x 250ml, ID 4x 536 g 4.5 4.5 8.0 30314821 Rotor, Angle, 6x 250ml FB, ID 6x 355 g 8.0 8.0 8.0 30314822 Rotor, Sw ing out, 4x 100ml, ID 4x 340 g 4.5 4.5 4.5 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 4x 465 g 4.0 5.0 5.0 4.5 4.5 30314824 Rotor, Sw ing out, 2x 3MTP, ID 4.5 2x 310 g 4.5 4.5 4.5 4.5 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 6x 140 g 11.0 13.5 9.0 9.0 30314826 Rotor, Angle, 6x 85ml RB, ID 11.0 6x 140 g 13.0 11.0 11.0 12.0 12.0 12.0 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 4x 140 g 12.0 15.0 15.0 4.5 30314828 Rotor, Sw ing out, 16x 50ml, ID 4x 557 g 4.5 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-35 7.5 7.5 30314829 Rotor, Angle, 10x 50ml FA, ID 10x 74 g 9.0 10.5 10.0 10.0 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 6x 72 g 6.0 6.0 6.0 12.0 12.0 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 6x 94 g 13.0 13.0 13.0 13.0 4.5 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 4.5 30x 32 g 4.5 4.5 4.5 12.0 12.0 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 20x 18 g 12.0 12.0 12.0 12.0 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 12x 25 g 6.0 6.0 6.0 13.5 13.5 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 15.0 44x 3.8 g 16.0 16.0 16.0 12.0 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 13.0 30x 3.4 g 14.0 15.0 15.0 14.0 14.0 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 15.0 24x 3.4 g 15.0 16.0 15.0 15.0 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 12x 3.4 g 18.0 30314840 Rotor, Angle, 64x 0.5ml, ID 64x 2.3 g 13.5 18.0 13.5 15.0 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 4x 3.5 g 15.0 15.0 15.0 30553085 Rotor Sw ingOut 4x 750ml D99mm ID SL 4x 995 g 4.0 30553086 Rotor Angle 4x 500ml D69mm ID 4x 708 g 8.0 4.0 8.0 TM DE-36 Frontier Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 1.2 Table 2: NIEDRIGSTE TEMPERATUREN UND MAX.-GESCHWINDIGKEIT Rotor-Nummer Rotorenbeschreibung 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314821 30314822 Rotor, Angle, 6x 250ml FB, ID Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 30314840 Rotor, Angle, 64x 0.5ml, ID 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID Modell FC5718R FC5816R FC5916R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5916R FC5916R Maximale Geschwindigkeit 15000 rpm 15000 rpm 16000 rpm 8000 rpm 4500 rpm 4500 rpm 5000 rpm 5000 rpm 4500 rpm 4500 rpm 4500 rpm 13500 rpm 13500 rpm 9000 rpm 9000 rpm 13000 rpm 12000 rpm 12000 rpm 16000 rpm 7500 rpm 7500 rpm 10500 rpm 6000 rpm 6000 rpm 12000 rpm 12000 rpm 13000 rpm 4500 rpm 4500 rpm 12000 rpm 12000 rpm 12000 rpm 6000 rpm 6000 rpm 13500 rpm 13500 rpm 16000 rpm 14000 rpm 14000 rpm 15000 rpm 15000 rpm 15000 rpm 16000 rpm 18000 rpm 18000 rpm 13500 rpm 13500 rpm 15000 rpm 15000 rpm 15000 rpm 4500 rpm 8000 rpm N-max 6 °C 6 °C 1 °C -1 ° C -7 ° C -7 ° C -3 ° C -3 ° C -6 ° C -6 ° C -8 ° C 17 ° C 17 ° C -5 ° C -5 ° C 8 °C 5 °C 5 °C 7 °C 9 °C 9 °C 2 °C -8 ° C -8 ° C -5 ° C -5 ° C -4 ° C -9 ° C -9 ° C 2 °C 2 °C -2 ° C -11.5 ° C -11.5 ° C 1 °C 1 °C 3 °C 9 °C 9 °C 3 °C 6 °C 6 °C 1 °C -2 ° C -2 ° C 2 °C 2 °C -1 ° C -1 ° C -2 ° C 4 °C 0 °C Alle Temperaturangaben beziehen sich auf eine Raumtemperatur von 23 ° C. Durch Überschreiten dieses Wertes oder direkte Sonneneinstrahlung auf die Zentrifuge können diese Werte nicht eingehalten werden. FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-37 1.3 Table 3: MAX. GESCHWINDIGKEIT UND RCF-WERTE FÜR ZULÄSSIGE ROTOREN Rotor-Nummer 30304361 30314820 30314821 30314822 30314823 30314824 30314825 30314826 Rotorenbeschreibung Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable Rotor, Sw ing out, 4x 250ml, ID Rotor, Angle, 6x 250ml FB, ID Rotor, Sw ing out, 4x 100ml, ID Rotor, Sw ing out, 4x 100ml, ID, Sealable Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID Modell Maximale Geschwindigkeit U/MIN RCF FC5714 14000 18624 x g FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5816 4500 3780 x g FC5816R 4500 3780 x g FC5816 8000 10016 x g FC5816R 8000 10016 x g FC5916 8000 10016 x g FC5916R 8000 10016 x g FC5714 4500 3350 x g FC5718 4500 3350 x g FC5718R 4500 3350 x g FC5714 4000 2486 x g FC5718 5000 3885 x g FC5718R 5000 3885 x g FC5714 4500 2716 x g FC5718 4500 2716 x g FC5718R 4500 2716 x g FC5816 4500 2716 x g FC5816R 4500 2716 x g FC5916 4500 2716 x g FC5916R 4500 2716 x g FC5718 11000 13932 x g FC5718R 13500 20984 x g FC5718 9000 10413 x g FC5718R 9000 10413 x g TM DE-38 Frontier Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314827 30314828 30314829 30314830 30314831 30314832 30314833 Rotor, Angle, 4x 85ml RB, ID, Hi Rotor, Sw ing out, 16x 50ml, ID Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Rotor, Angle, 20x 10ml RB, ID, Hi FC5816 11000 15555 x g FC5816R 13000 21726 x g FC5916 11000 15555 x g FC5916R 13000 21726 x g FC5718 12000 14809 x g FC5718R 12000 14809 x g FC5816 12000 14809 x g FC5816R 12000 14809 x g FC5916 15000 23140 x g FC5916R 16000 26328 x g FC5816 4500 3735 x g FC5816R 4500 3735 x g FC5718 7500 8174 x g FC5718R 7500 8174 x g FC5816 9000 11771 x g FC5816R 10500 16022 x g FC5916 10000 14532 x g FC5916R 10500 16022 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 12000 13522 x g FC5718R 12000 13522 x g FC5816 13000 15869 x g FC5816R 13000 15869 x g FC5916 13000 15869 x g FC5916R 13000 15869 x g FC5714 4500 2830 x g FC5718 4500 2830 x g FC5718R 4500 2830 x g FC5816 4500 2830 x g FC5816R 4500 2830 x g FC5718 12000 15775 x g FC5718R 12000 15775 x g FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-39 30314834 30314835 30314836 30314838 30314839 30314840 30314841 Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable Rotor, Angle, 24x 1.5/2.0ml, ID Rotor, Angle, 12x 1.5/2.0ml, ID Rotor, Angle, 64x 0.5ml, ID Rotor, Angle, 4x 8-w PCR Strip, ID FC5816 12000 15775 x g FC5816R 12000 15775 x g FC5916 12000 15775 x g FC5916R 12000 15775 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 13500 17113 x g FC5718R 13500 17113 x g FC5816 15000 21128 x g FC5816R 16000 24039 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5714 12000 15131 x g FC5718 13000 17758 x g FC5718R 14000 20595 x g FC5916 15000 23643 x g FC5916R 15000 23643 x g FC5714 14000 18624 x g FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5718 18000 23542 x g FC5718R 18000 23542 x g FC5718 13500 16298 x g FC5718R 13500 16298 x g FC5718 15000 15343 x g FC5718R 15000 15343 x g FC5916 15000 15343 x g FC5916R 15000 15343 x g DE-40 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID FC5916 4000 3434 x g FC5916R 4500 4346 x g FC5916 8000 10367 x g FC5916R 8000 10367 x g FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-41 11.4 Table 4: BESCHLEUNIGUNGS- UND VERLANGSAMUNGSZEITEN Artikelnummer Rotorbeschreibung Modell Beschleunigung* Zeit in sek level 0 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314820 Rotor, Sw ing out, 4x 250ml, ID 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314828 Rotor, Sw ing out, 16x 50ml, ID 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R 200 240 230 251 249 204 204 309 309 664 664 573 573 110 100 100 110 100 100 230 160 160 452 452 249 249 400 500 360 360 697 2313 463 549 300 300 506 506 448 448 311 311 380 380 753 753 480 480 101 89 90 level 9 25 23 25 25 20 21 21 34 34 130 130 66 66 14 15 15 14 15 15 25 20 20 43 43 27 27 40 60 40 40 85 70 48 69 60 60 60 60 50 50 34 34 63 62 115 115 60 60 11 13 12 Abbremsung* Zeit in sek level 0 230 240 420 610 635 421 421 458 458 2906 2906 1903 1903 170 150 150 170 150 150 340 360 360 616 616 488 488 960 1260 1050 1050 2313 1630 1654 1307 820 820 1745 1745 1251 1251 387 387 796 1069 2395 2395 1747 1747 206 239 463 level 9 35 16 17 26 27 30 30 36 36 92 92 84 84 19 15 15 17 15 15 26 15 15 38 38 23 23 40 50 40 40 70 76 46 67 40 40 49 49 45 45 36 36 37 36 72 72 68 68 14 11 11 DE-42 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 30314840 Rotor, Angle, 64x 0.5ml, ID 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 30553085 Rotor Sw ingOut 4x 750ml D99mm ID SL 30553086 Rotor Angle 4x 500ml D69mm ID FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5916 FC5916R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5718 FC5718R FC5916 FC5916R FC5916 FC5916R FC5916 FC5916R * Hinweis: Beschleunigt von 0 auf Vmax; bremst von Vmax auf 0 ab. 360 360 446 446 264 264 157 160 160 149 149 360 360 512 512 305 305 102 88 88 196 196 257 274 233 233 250 200 210 221 221 200 210 230 251 249 204 204 240 240 150 150 130 130 100 100 483 551 575 575 40 40 48 48 28 28 19 15 15 25 25 40 40 54 54 32 32 11 13 12 22 22 29 31 25 25 27 25 25 23 23 22 23 25 25 30 21 21 25 25 17 17 16 16 12 12 47 55 73 73 570 570 1323 1323 921 921 370 380 380 985 985 570 570 1439 1439 988 988 167 222 433 204 204 436 485 283 500 280 360 360 561 561 230 240 420 610 635 421 421 210 210 170 170 130 130 201 201 1287 1501 2317 2317 30 30 49 49 32 32 20 10 10 20 20 30 30 47 47 37 37 14 11 11 18 18 28 29 30 25 34 25 30 30 30 35 16 17 26 27 30 30 20 20 12 12 12 12 12 12 49 54 82 82 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-43 11.5 Tabelle 5: FEHLERMITTEILUNGEN Error-No.: Descri pti on 1 I mbal ance arose 2 I mbal ance sensor i s defecti v e 4 I mbal ance swi tch has been acti vated for l onger than 5 secon ds 8 T ransponder i n the rotor i s defecti v e 11 T emperatu r e sensor i s defecti v e 12 Chamber over temperature 14 Leap of speed i s too bi g between two mesaurem e nts CLOSE l i d 33 34 38 40 Open l i d whi l e motor i s runni ng Li d contact defecti v e Li d motor i s bl ocked Communi cati o n wi th frequen cy converter di strubed duri ng start 41 Communi cati o n wi th frequen cy converter di strubed duri ng stop 42 Short ci rcui t i n the frequency converter 43 Undervol tag e frequen cy converter 44 Overvol tag e frequen cy converter 45 Over temperatu r e frequency converter 46 Over temperatu r e motor 47 Over current frequen cy converter 48 T i meout between control uni t and frequency converter 49 Other error frequency conver ter 55 Overspeed 70 T i meout between control er and RS232 i nterface 99 Rotor i s not al l owed i n thi s centri fuge FALSE rotor no I nserted rotor does not exi st i n the program m Rotor i s not detected TM DE-44 Frontier Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.6 Table 6: RADIUSKORREKTUR FC5916 FC5916R Korrektur (cm) FC5816R Rotor, Angle, 6x 250ml FB, ID Radius (cm) FC5816 30314821 Rotor, Sw ing out, 4x 250ml, ID Buchse/Adapter Beschreibung FC5718R 30314820 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable Buchse/Ad apter FC5718 30304361 Rotorbeschreibun g FC5714 Rotor Kompatibel ● ● ● ● ● ● ● -0.3 ● ● ● ● ● ● ● -1.0 ● ● ● ● ● ● ● ● ● -1.0 ● ● 15.9 -0.8 ● ● Rack, 3x 50ml D29mm RB, 2/pk 16.3 -0.4 ● ● 30314906 Rack, 3x 50ml D29mm FA, 2/pk 16.7 0.0 ● ● 30314907 Rack, 1x 50ml D34mm RB, 2/pk 15.8 -0.9 ● ● 30314908 Rack, 2x 50ml D29mm Rim, 2/pk 15.8 -0.9 ● ● 30314909 Rack, 1x 50ml D29mm Rim, 2/pk 16.1 -0.6 ● ● 30314910 Rack, 5x 25ml D24.5mm RB, 2/pk 16.1 -0.6 ● ● 30314911 Rack, 3x 25ml D24.5mm Rim, 2/pk 15.5 -1.2 ● ● 30314912 Rack, 9x 15ml D17mm RB, 2/pk 16.3 -0.4 ● ● 30314913 Rack, 7x 15ml D17mm FA, 2/pk 16.4 -0.3 ● ● 30314914 Rack, 7x 15ml D17mm Rim, 2/pk 16.1 -0.6 ● ● 30314915 Rack, 8x 10ml D16mm RB, 2/pk 16.3 -0.4 ● ● 30314916 Rack, 14x 5-7ml D13mm RB, 2/pk 15.9 -0.8 ● ● 30314917 Rack, 10x 1.6-7ml D13mm Vac, 2/pk 15.9 -0.8 ● ● 30304367 Rack, 8x 4-10ml D16mm Vac, 2/pk 16.3 -0.4 ● ● 30304368 Adapter, 9x 1.5/2.0ml D11mm, 2/pk 15.7 -1.0 ● ● Blank NO Bucket/Adapter 14.1 ● ● ● ● 30304369 Adapter, 8x 1.5ml D11mm, 2/pk 13.2 -0.9 ● ● ● ● 30304370 Adapter, 5x 10ml D16mm RB, 2/pk 13.3 -0.8 ● ● ● ● 30304371 Adapter, 4x 15ml D17mm FA, 2/pk 13.0 -1.1 ● ● ● ● 30304372 Adapter, 2x 30ml D26mm RB, 2/pk 12.5 -1.6 ● ● ● ● 30304373 Adapter, 1x 50ml D29mm RB, 2/pk 12.0 -2.1 ● ● ● ● Blank NO Bucket/Adapter 8.5 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 Blank NO Bucket/Adapter 16.7 30314903 Rack, 1x 250ml D62mm FB, 2/pk 15.7 30314904 Rack, 1x 100ml D41mm RB, 2/pk 30314905 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-45 30314822 30314823 Rotor, Sw ing out, 4x 100ml, ID Rotor, Sw ing out, 4x 100ml, ID, Sealable 30304374 Adapter, 1x 50ml D30mm FA, 2/pk 11.7 30559412 Adapter 15x 1.5/2.0ml D11mm MTBx 2 V1 9.5 13.6 30559413 Adapter 1x 250ml D60mm CBCx 3 V1 11.8 30559414 Adapter 1x 175/225ml D61mm FAx 2 V1 12.8 Blank NO Bucket/Adapter 14.8 30314842 Rack, 1x 100ml D46mm RB, 2/pk 14.1 30314843 Rack, 1x 85ml D38mm RB, 2/pk 30314844 -2.4 ● ● ● ● -4.6 -0.5 ● ● ● ● -0.5 ● ● ● ● ● ● ● ● -2.3 ● ● ● -0.7 ● ● ● 14.1 -0.7 ● ● ● Rack, 1x 50ml D34mm RB, 2/pk 14.2 -0.6 ● ● ● 30314845 Rack, 1x 50ml D29mm RB, 2/pk 14.2 -0.6 ● ● ● 30314846 Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.7 ● ● ● 30314847 Rack, 1x 50ml D29mm Rim, 2/pk 14.3 -0.5 ● ● ● 30314848 Rack, 1x 30ml D25mm Rim, 2/pk 13.8 -1.0 ● ● ● 30314849 Rack, 1x 25ml D25mm RB, 2/pk 14.4 -0.4 ● ● ● 30314850 Rack, 2x 15ml D17mm FA, 2/pk 14.8 0.0 ● ● ● 30314851 Rack, 1x 15ml D17mm Rim, 2/pk 14.5 -0.3 ● ● ● 30314852 Rack, 7x 15ml D17mm RB, 2/pk 14.5 -0.3 ● ● ● 30314853 Rack, 9x 5-7ml D13mm RB, 2/pk 14.2 -0.6 ● ● ● 30314854 Rack, 10x 1.5ml D11mm, 2/pk 9.9 -4.9 ● ● ● 30314855 Rack, 9x 4.5ml D12mm FB, 2/pk 14.0 -0.8 ● ● ● 30314856 Rack, 7x 5-7ml D13mm RB, 2/pk 11.6 -3.2 ● ● ● 30314857 Rack, 7x 5-7ml D13mm Vac, 2/pk 14.2 -0.6 ● ● ● 30314858 Rack, 5x 4-10ml D16mm Sar, 2/pk 14.3 -0.5 ● ● ● Blank NO Bucket/Adapter 14.6 ● ● ● 30314860 Bucket, 100ml w o Cap, Sealable, 2/pk 14.2 -0.4 ● ● ● 30314861 Bucket, 100ml w / Cap, Sealable, 2/pk 14.2 -0.4 ● ● ● 30314864 Rack, 1x 100ml D40mm RB, 2/pk 13.8 -0.8 ● ● ● 30314865 Rack, 1x 85ml D38mm RB, 2/pk 14.0 -0.6 ● ● ● 30314866 Rack, 1x 50ml D34mm RB, 2/pk 14.0 -0.6 ● ● ● 30314867 Rack, 1x 50ml D29mm RB, 2/pk 14.0 -0.6 ● ● ● 30314868 Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.5 ● ● ● 30314869 Rack, 1x 50ml D29mm Rim, 2/pk 13.9 -0.7 ● ● ● DE-46 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314824 30314825 30314826 Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID 30314870 Rack, 1x 30ml D25mm Rim, 2/pk 13.1 -1.5 ● ● ● 30314871 Rack, 1x 25ml D25mm RB, 2/pk 14.0 -0.6 ● ● ● 30314872 Rack, 1x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314873 Rack, 2x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314874 Rack, 2x 15ml D17mm Rim, 2/pk 14.0 -0.6 ● ● ● 30314875 Rack, 4x 15ml D17mm RB, 2/pk 14.0 -0.6 ● ● ● 30314876 Rack, 7x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314877 Rack, 5x 1.5ml D11mm, 2/pk 14.0 -0.6 ● ● ● 30314878 Rack, 3x 10ml D17mm Sar, 2/pk 14.0 -0.6 ● ● ● 30314879 Rack, 4x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314880 Rack, 4x 10ml D17mm Vac, 2/pk 14.0 -0.6 ● ● ● 30314881 Bucket, 2x 50ml D29mm FA, w / Rack, 2/pk 14.6 0.0 ● ● ● 30314882 Bucket, 3x 15ml D17mm FA, w / Rack, 2/pk 14.6 0.0 ● ● ● 30314883 Cy to, Rectangular Bucket, 2/pk 6.3 -8.3 ● ● ● Blank NO Bucket/Adapter 12.0 ● ● ● ● ● ● ● 30314891 Adapter, 48x 1.5/2.0ml D11mm, 2/pk 12.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 10.3 ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 9.6 -0.7 ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 9.6 -0.7 ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 10.0 -0.3 ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 9.8 -0.5 ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 9.3 -1.0 ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 10.3 0.0 ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 9.5 -0.8 ● ● Blank NO Bucket/Adapter 11.5 ● ● ● ● ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 10.9 -0.6 ● ● ● ● ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 11.1 -0.4 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● 0.0 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-47 30314827 30314828 30314829 30314830 30314831 30314832 Rotor, Angle, 4x 85ml RB, ID, Hi Rotor, Sw ing out, 16x 50ml, ID Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Blank NO Bucket/Adapter 9.2 ● ● ● ● ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 8.5 -0.7 ● ● ● ● ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 8.4 -0.8 ● ● ● ● ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 8.9 -0.3 ● ● ● ● ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 8.6 -0.6 ● ● ● ● ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 8.3 -0.9 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 7.5 -1.7 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 8.3 -0.9 ● ● ● ● ● ● Blank NO Bucket/Adapter 16.5 ● ● 30304375 Rack, 4x 50ml D29mm FA, 2/pk 16.5 0.0 ● ● 30314583 Rack, 10x 15ml D17mm FA, 2/pk 16.5 0.0 ● ● Blank NO Bucket/Adapter 13.0 30130894 Adapter, 1x 15ml D17mm RB, 2/pk 12.8 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk Blank NO Bucket/Adapter 11.0 30130891 Adapter, 1x 30ml D26mm, 2/pk 10.7 30130892 Adapter, 1x 16ml D18mm, 2/pk 30130893 ● ● ● ● ● ● -0.2 ● ● ● ● ● ● 12.2 -0.8 ● ● ● ● ● ● 10.4 -2.6 ● ● ● ● ● ● 8.9 -4.1 ● ● ● ● ● ● ● ● ● -0.3 ● ● ● 10.3 -0.7 ● ● ● Adapter, 1x 15ml D17mm, 2/pk 10.6 -0.4 ● ● ● 30130894 Adapter, 1x 15ml D17mm RB, 2/pk 10.6 -0.4 ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 10.2 -0.8 ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 8.3 -2.7 ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 6.7 -4.3 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130891 Adapter, 1x 30ml D26mm, 2/pk 8.2 -0.2 ● ● ● ● ● ● 30130892 Adapter, 1x 16ml D18mm, 2/pk 7.9 -0.5 ● ● ● ● ● ● 30130893 Adapter, 1x 15ml D17mm, 2/pk 8.0 -0.4 ● ● ● ● ● ● 30314892 Adapter, 1x 15ml D17.5mm FA, 2/pk 7.7 -0.7 ● ● ● ● ● ● Blank NO Bucket/Adapter 12.5 ● ● ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 12.2 -0.3 ● ● ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 10.5 -2.0 ● ● ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 9.0 -3.5 ● ● ● ● ● DE-48 FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314834 30314835 30314836 30314838 30314839 30553085 Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable Rotor, Angle, 24x 1.5/2.0ml, ID Rotor, Angle, 12x 1.5/2.0ml, ID Rotor Sw ingOut 4x 750ml D99mm ID SL Blank NO Bucket/Adapter 11.0 ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 10.6 -0.4 ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 9.1 -1.9 ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 7.7 -3.4 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.2 ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.7 -0.7 ● ● ● ● ● ● Blank NO Bucket/Adapter 9.4 ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 9.1 -0.3 ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 8.4 -1.0 ● ● ● ● ● Blank NO Bucket/Adapter 8.5 ● ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.3 ● ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 -1.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 6.5 ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 6.3 -0.2 ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 5.6 -0.9 ● ● 30314900 Adapter, 1x 1.5ml for ov er16000x g, 6/pk 6.4 -0.1 ● ● Blank NO Bucket/Adapter 19.3 ● ● 30553138 Adapter 32x 7ml D13mm RBx 2 V1 18.7 -0.6 ● ● 30553135 Adapter 18x 15ml D17mm RBx 2 V1 18.8 -0.5 ● ● 30553132 Adapter 8x 30ml D26mm RBx 2 V1 19.1 -0.2 ● ● 30553131 Adapter 7x 50ml D29mm RBx 2 V1 19.1 -0.2 ● ● 30553127 Adapter 3x 100ml D41mm RBx 2 V1 19.1 -0.2 ● ● 30553125 Adapter 1x 250ml D62mm FBx 2 V1 18.9 -0.4 ● ● 30553126 Adapter 1x 225/190ml D61mm FAx 2 V1 19.1 -0.2 ● ● 30553128 Adapter 5x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553130 Adapter 7x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553133 Adapter 14x 15ml D17mm FAx 2 V1 19.2 -0.1 ● ● 30553134 Adapter 17x 15ml D17mm FAx 2 V1 19.0 -0.3 ● ● 30553140 Adapter 18x 5/10ml D16mm Vac/Sarx 2 V1 19.0 -0.3 ● ● 30559377 Adapter 38x 1.1/1.4ml D8mm Sarx 2 V1 18.9 -0.4 ● ● 30553136 Adapter 21x 10ml D16mm RBx 2 V1 18.9 -0.4 ● ● FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R DE-49 30553086 Rotor Angle 4x 500ml D69mm ID 30553139 Adapter 27x 1.8/7ml D13mm Vac/Sarx 2 V1 18.8 Blank NO Bucket/Adapter 14.5 30564850 Adapter 1x 250ml D61.5mm FBx 2 V1 13.7 30559417 Adapter 1x 175/225ml D61.5mm FAx 2 V1 30559419 -0.5 ● ● ● ● -0.8 ● ● 13.4 -1.1 ● ● Adapter 1x 50ml D29.5mm FAx 2 V1 12.4 -2.1 ● ● 30559416 Adapter 1x 250ml D60.5mm CBCx 2 V1 12.6 -1.9 ● ● 30559421 Adapter 4x 30ml D25.5mm RBx 2 V1 13.8 -0.7 ● ● 30559420 Adapter 3x 50ml D29mm RBx 2 V1 14.3 -0.2 ● ● 30559422 Adapter 6x 15ml D17mm FAx 2 V1 13.8 -0.7 ● ● FrontierTM Zentrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.7 Tabelle 7: Rüc knahmeformular / Dekontaminierungszertifikat Fügen Sie dieses Formular allen Rücksendungen von Ausrüstungs- und Aufbauteilen bei! Die ausgefül lte Erklärung über die Dekontamination ist eine Vorkehrung für di Annahme und die weitere Bearbeitung der Rücksendung. Wenn keine entsprechend Erläuterung beigefüg t ist, führen wir die Dekontamination auf Ihre Kosen durch. Nachname; Familienname: Bitte in Blockbuchstaben ausfüllen!  DE-50 Organisation / Unternehmen: Straße: Postfach: Ort: Telefon: Fax: E-Mail: Pos. Crow d De contam inate d obje ct Se rial num be r De scription / Com m e nt 1 2 3 4 Sind die oben aufgeführten Teile in Kontakt mit den folgenden Substanzen? Gesundheitsgefährdende wässrige Lösungen, Puffer, Säuren, Basen:…………………….....  Ja  Nein Potentielle ansteckende Mittel: ……………………………………………………………  Ja  Nein Organische Reagenzien und Lösungsmittel: ……………………………………………………….. .. Ja  Nein Radioaktive Substanzen: ……………………………………………. α.. β.. γ…  Ja  Nein Gesundheitsgefährdende Proteine: …………………………………………………………..  Ja  Nein DNA: ………………………………………………………………………………………  Ja  Nein Diese Substanzen haben die Ausrüstung/den Aufbau erreicht? ………………………. Welche, falls ja: Beschreibung der Maßnahmen zur Dekontamination der aufgeführten Teile: Ich bestätige die ordnungsgemäße Dekontaminationn: Unternehmen/Abt . Ort und Datum:  Ja  Nein Schizzo della mappa di FC5714 / FC5718 / FC5816 / FC5916 1 2 7 4 3 6 5 Figura 1 Schizzo della mappa di FC5718R 1 2 7 3 4 6 5 Figura 2 Schizzo della mappa di FC5816R / FC5916R 1 2 7 3 6 4 5 Figura 3 1 Coperchio Centrifuga 3 Display 5 Interruttore Principale di Alimentazione 7 Rilascio di Emergenza 2 Finestra Rotore 4 Etichetta Funzionamento 6 Connessione di Alimentazione Etichetta di funzionamento per FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figura 4 1 4 5 6 7 8 2 3 9 10 11 12 Figura 5 1. 3. 5. 7. 9. Display LCD Avvio centrifuga Modello e selezione intensità Accelerazione/Decelerazione Modello impostazione tempo Centrifuga a rotazione lenta/veloce 11. Arresto/Impostazioni centrifuga 2. 4. 6. Modello impostazione programma Modello e selezione RPM/RCF Modello impostazione temperatura (Solo FC5515R) 8. Salvare le informazioni di impostazione 10. Coperchio di rilascio 12. Manopola/Quadrante di regolazione: Modificare il numero Display LCD L'immagine qui di seguito illustra i singoli elementi del display LCD Figura 6 M2 M3 M1 M12 M15 M13 M4 M9 M11 M10 M5 M6 M7 A-1 M8 A-2 M14 A-3 A-4 Figura 7 Campi display: A-1 A-2 A-3 A-4 Campo display – "RPM/RCF" Campo display – "Acc/Dec" "Service" Campo display – "Time/Prog" Campo display – "Temp" Messaggi/logo dei campi display M1 M4 M7 M10 M13 "chiudi" "Rotore-N°." "accel" "programma" "h m s" M2 M5 M8 M11 "apri" "rpm" "decel" "errore" M3 M6 M9 M12 M14 "rotore" "rcf" "raggio" "servizio" "temperatura" M15 "preraffreddamento" Tabella Informazioni Rotore 24 25 26 27 28 29 30 31 32 33 34 35 36 30314824 30314825 30314826 30314827 30314828 30314829 30314830 30314831 30314832 30314833 30314834 30314835 30314836 38 39 40 41 85 30314838 30314839 30314840 30314841 30553085 86 30553086 ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● FC5916R 30314820 30314821 30314822 30314823 ● FC5916 20 21 22 23 Rotor, Angle, 24x1.5/2.0ml, ID, Sealable Rotor, Swing out, 4x250ml, ID Rotor, Angle, 6x250ml FB, ID Rotor, Swing out, 4x100ml, ID Rotor, Swing out, 4x100ml, ID, Sealable Rotor, Swing out, 2x3MTP, ID Rotor, Angle, 6x85ml RB, ID, Hi Rotor, Angle, 6x85ml RB, ID Rotor, Angle, 4x85ml RB, ID, Hi Rotor, Swing out, 16x50ml, ID Rotor, Angle, 10x50ml FA, ID Rotor, Angle, 6x50ml RB/FA, ID Rotor, Angle, 6x50ml RB, ID, Hi Rotor, Angle, 30x15ml RB/FA, ID Rotor, Angle, 20x10ml RB, ID, Hi Rotor, Angle, 12x15ml RB/FA, ID Rotor, Angle, 44x1.5/2.0ml, ID Rotor, Angle, 30x1.5/2.0ml, ID, Sealable Rotor, Angle, 24x1.5/2.0ml, ID Rotor, Angle, 12x1.5/2.0ml, ID Rotor, Angle, 64x0.5ml, ID Rotor, Angle, 4x8-w PCR Strip, ID Rotor SwingOut 4x750ml D99mm ID SL Rotor Angle 4x500ml D69mm ID FC5816R 30304361 FC5816 61 Compatibile FC5718R Capacità FC5718 Numero d'ordine FC5714 Visualizzazi one numero del rotore ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● 1. 2. 3. 4. 5. SOMMARIO INTRODUZIONE ............................................................................................................................................................. 1 1.1 DESCRIZIONE E SCOPO ........................................................................................................................................................ 1 1.2 BREVE DESCRIZIONE ............................................................................................................................................................ 1 1.3 DEFINIZIONE DEI SEGNALI E DEI SIMBOLI DI AVVERTENZA ................................................................................................. 1 1.4 PRECAUZIONI DI SICUREZZA ................................................................................................................................................ 2 1.4.1 UTENTE ........................................................................................................................................................................ 2 1.4.2 ROTORE E ACCESSORI .................................................................................................................................................. 2 1.4.3 PROVVEDIMENTI PER LA VOSTRA SICUREZZA ............................................................................................................ 2 1.4.4 ESCLUDERE LE INFLUENZE AMBIENTALI SEGUENTI ..................................................................................................... 2 1.4.5 PROVVEDIMENTI PER LA SICUREZZA DURANTE IL FUNZIONAMENTO ........................................................................ 3 1.4.6 PERICOLI E PRECAUZIONI ............................................................................................................................................. 3 1.4.7 ABBREVIAZIONI UTILIZZATE NEL PRESENTE MANUALE ............................................................................................... 3 INSTALLAZIONE ............................................................................................................................................................. 4 2.1 DISIMBALLO ......................................................................................................................................................................... 4 2.1.1 IMBALLAGGIO DI CONSEGNA ...................................................................................................................................... 4 2.2 SELEZIONARE L'AMBIENTE .................................................................................................................................................. 4 2.3 INSTALLAZIONE .................................................................................................................................................................... 4 2.4 PRECAUZIONI DI SICUREZZA DURANTE IL FUNZIONAMENTO E GARANZIA ........................................................................ 5 FUNZIONAMENTO ......................................................................................................................................................... 5 3.1 MONTAGGIO E CARICAMENTO DEL ROTORE ...................................................................................................................... 5 3.1.1 INSTALLAZIONE DEI ROTORI ........................................................................................................................................ 5 3.1.2 CARICO DEI ROTORI D'ANGOLO ................................................................................................................................... 6 3.1.3 CARICO DEI ROTORI A BATTENTE ................................................................................................................................ 6 3.1.4 CARICO E SOVRACCARICO DEI ROTORI ........................................................................................................................ 7 3.1.5 RIMOZIONE DEL ROTORE ............................................................................................................................................. 7 3.2 CONTROLLO DEL COPERCHIO .............................................................................................................................................. 7 3.2.1 APERTURA DEL COPERCHIO ......................................................................................................................................... 7 3.2.2 BLOCCAGGIO DEL COPERCHIO..................................................................................................................................... 8 3.3 PRESELEZIONE ..................................................................................................................................................................... 8 3.3.1 PRESELEZIONE DELLA VELOCITÀ / VALORE RCF ........................................................................................................... 8 3.3.2 PRESELEZIONE DEL TEMPO DI FUNZIONAMENTO ....................................................................................................... 9 3.3.3 PRESELEZIONE DELL'INTENSITÀ DEL FRENO E DELL'ACCELERAZIONE ........................................................................ 9 3.3.4 PRESELEZIONE DELLA TEMPERATURA (SOLO MODELLI REFRIGERATI) ...................................................................... 10 3.3.5 PRE-RAFFREDDAMENTO (SOLO MODELLI REFRIGERATI).......................................................................................... 10 3.4 CORREZIONE DEL RAGGIO ................................................................................................................................................. 10 3.5 PROGRAMMA .................................................................................................................................................................... 11 3.5.1 SALVATAGGIO DI PROGRAMMI ................................................................................................................................. 11 3.5.2 RICHIAMO DEI PROGRAMMI SALVATI ....................................................................................................................... 12 3.5.3 LASCIARE LA MODALITÀ PROGRAMMA ..................................................................................................................... 12 3.6 AVVIAMENTO E ARRESTO DELLA CENTRIFUGA ................................................................................................................. 12 3.6.1 AVVIARE LA CENTRIFUGA .......................................................................................................................................... 12 3.6.2 IL TASTO "STOP" ........................................................................................................................................................ 13 3.7 RILEVAMENTO DI SQUILIBRI .............................................................................................................................................. 13 IMPOSTAZIONE ........................................................................................................................................................... 14 4.1 REGOLAZIONI DI BASE ....................................................................................................................................................... 14 4.1.1 ACCESSO ALLA MODALITÀ "DATI DI FUNZIONAMENTO" .......................................................................................... 14 4.1.2 INDICAZIONE DELLA TEMPERATURA ......................................................................................................................... 14 4.1.3 ACCENSIONE/SPEGNIMENTO SEGNALE ..................................................................................................................... 15 4.1.4 PRE-SELEZIONE DEL VOLUME DEL SEGNALE ACUSTICO ............................................................................................ 15 4.1.5 SELEZIONE DELLA SUONERIA DEL SEGNALE ACUSTICO "FINE CICLO" ...................................................................... 16 4.1.6 ACCENSIONE/SPEGNIMENTO DEL SUONO TASTIERA ................................................................................................ 16 4.1.7 RICHIAMO DEI DATI DI FUNZIONAMENTO ................................................................................................................ 17 MANUTENZIONE ......................................................................................................................................................... 17 5.1 MANUTENZIONE E PULIZIA................................................................................................................................................ 17 5.1.1 GENERALITÀ .............................................................................................................................................................. 17 5.1.2 PULIZIA E DISINFEZIONE DEL GRUPPO ...................................................................................................................... 18 5.1.3 PULIZIA E DISINFEZIONE DEL ROTORE ....................................................................................................................... 18 5.1.4 DISINFEZIONE DEI ROTORI IN ALLUMINIO ................................................................................................................. 18 5.1.5 DISINFEZIONE DEI ROTORI IN POLIPROPILENE .......................................................................................................... 19 5.1.6 ROTTURA DEL VETRO ................................................................................................................................................. 19 5.2 DURATA IN VITA DI ROTORI, CONTENITORI E ACCESSORI ................................................................................................. 19 6. RICERCA GUASTI .......................................................................................................................................................... 20 6.1 MESSAGGIO DI ERRORE: CAUSA/SOLUZIONE .................................................................................................................... 20 6.2 ESAME DEI POSSIBILI MESSAGGI DI ERRORE E DELLE RELATIVE SOLUZIONI ..................................................................... 20 6.2.1 RILASCIO DEL COPERCHIO DURANTE UNA CADUTA DI TENSIONE (RILASCIO DEL COPERCHIO DI EMERGENZA)...... 20 6.2.2 DESCRIZIONE DEL SISTEMA DEI MESSAGGI DI ERRORE ............................................................................................. 21 7. RICEZIONE DELLE CENTRIFUGHE PER RIPARAZIONE ..................................................................................................... 21 8. TRASPORTO, STOCCAGGIO E SMALTIMENTO ............................................................................................................... 22 8.1 TRASPORTO ....................................................................................................................................................................... 22 8.2 STOCCAGGIO ..................................................................................................................................................................... 22 8.3 TRASPORTO, INSTALLAZIONE E SMALTIMENTO DELLA CENTRIFUGA FC5714/FC5718/FC5718R/FC5816/FC5816R ....... 22 8.3.1 TRASPORTO................................................................................................................................................................ 22 8.3.2 INSTALLAZIONE .......................................................................................................................................................... 22 8.3.3 IMBALLAGGIO ............................................................................................................................................................ 23 8.3.4 TRASFERIMENTO DEL DISPOSITIVO ........................................................................................................................... 23 9. DATI TECNICI ............................................................................................................................................................... 24 9.1 SPECIFICHE ......................................................................................................................................................................... 24 9.1.1 CENTRIFUGA FC5714................................................................................................................................................. 24 9.1.2 CENTRIFUGA FC571................................................................................................................................................... 25 9.1.3 CENTRIFUGA FC5718R ............................................................................................................................................... 26 9.1.4 CENTRIFUGA FC5816 ................................................................................................................................................. 27 9.1.5 CENTRIFUGA FC5816R ............................................................................................................................................... 28 9.2 DISEGNI E DIMENSIONI...................................................................................................................................................... 29 10. CONFORMITA .............................................................................................................................................................. 31 11. APPENDICE .................................................................................................................................................................. 32 11.1 TABELLA 1: PESO NETTO AMMESSO................................................................................................................................ 33 11.2 TABELLA 2: TEMPERATURE PIÙ BASSE ALLA MAX. VELOCITÀ (SOLO MODELLI REFRIGERATI) ....................................... 34 11.3 TABELLA 3: VELOCITÀ MAX. E VALORI RCF PER ROTORI AMMISSIBILI ............................................................................ 35 11.4 TABELLA 4: TEMPI DI ACCELERAZIONE E DECELERAZIONE .............................................................................................. 37 11.5 TABELLA 5: MESSAGGI DI ERRORE ................................................................................................................................... 41 11.6 TABELLA 6: SPECIFICHE CORREZIONE RAGGIO E ADATTATORE ...................................................................................... 42 11.7 TABELLA 7: MODULO DI RISANAMENTO/CERTIFICATO DI DECONTAMINAZIONE .......................................................... 46 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-1 1. INTRODUZIONE 1.1 Descrizione e Scopo Grazie per aver scelto questo prodotto OHAUS Tutti i simboli indicano istruzioni di sicurezza e punti di situazioni potenzialmente pericolosi. Vi invitiamo a leggere TM interamente il manuale prima di utilizzare Frontier MultiPro al fine di evitare funzionamenti errati. TM Le centrifughe Frontier MultiPro sono state progettate per la separazione di materiali o miscele con differenti densità. Le centrifughe OHAUS sono intese esclusivamente per utilizzo all'interno e da parte di personale qualificato. 1.2 Breve descrizione I modelli FC5714/FC5718/FC5816/FC5916 sono centrifughe universali non refrigerate. I modelli FC5718R/FC5816R/FC5916R sono centrifughe universali refrigerate. Tutti e tre i modelli sono offerti in due variazioni di tensione, 230V o 120V. Le centrifughe possono essere utilizzati con rotori a battenti e ad angolo Tutti i parametri sono accessibili mediante pulsanti e selezionati mediante il regolatore centrale. Tutti i valori preselezionati e attuali saranno mostrati in modo permanente sul display LCD. La centrifuga è alimentata da un motore a induzione che non richiede manutenzione. I dati tecnici dettagliati sono contenuti nella sezione "Dati Tecnici". 1.3 Definizione dei Segnali e dei Simboli di Avvertenza Le avvertenze di sicurezza son contrassegnate con diciture e simboli di avvertenza Essi indicano problemi riguardanti la sicurezza e le relative avvertenze. La mancata attenzione alle avvertenze di sicurezza può provocare lesioni personali, danni allo strumento, malfunzionamenti e falsi risultati. Il grado di pericolosità è una parte dell'avvertenza di sicurezza e distingue l'uno dall'altro i possibili risultati della mancata osservanza . Diciture PERICOLO AVVERTENZA PRECAUZIONE ATTENZIONE NOTA Se non evitato, provocherà lesioni gravi o morte. Per una situazione pericolosa con rischio medio., che può provocare lesioni o more se non evitata. Per una situazione pericolosa con rischio ridotto, che può produrre danni alla proprietà oppure perdita di dati o lesione se non evitata. Per informazioni importanti riguardanti il prodotto Se non evitate, possono provocare danni all'apparecchiatura Per informazioni utili riguardanti il prodotto Simboli di Avvertenza Pericolo Generico Pericolo di Scossa Elettrica Corrente Alternata Rischio Biologico Esplosione Schiacciamento IT-2 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Avvertenza e segnali di informazione sulla superficie della centrifuga Occorre utilizzare sempre quattro supporti sui rotori a battente a quattro punti, altrimenti la centrifuga subirà dei danni. Detti danni non saranno coperti dalla garanzia sul prodotto. Attenzione! Verificare il serraggio del dado del rotore prima di ogni funzionamento. Staccare la spina di alimentazione di rete prima di aprire la sede o il rilascio di emergenza. Senso di rotazione - rotazione oraria per la motorizzazione del rotore Riferimenti per il carico dei rotori 1.4 Precauzioni di sicurezza 1.4.1 Utente Le centrifughe OHAUS sono intese esclusivamente per utilizzo all'interno e da parte di personale qualificato. Il dispositivo può essere utilizzato solo da personale specializzato e formato. Detto personale deve aver letto attentamente il manuale di funzionamento e avere familiarità con il funzionamento del dispositivo. 1.4.2 Rotore e accessori Si devono utilizzare solo rotori e accessori originali OHAUS. Ogni altro utilizzo o utilizzo inteso deve essere considerato inappropriato. OHAUS non è responsabile per danni derivanti dall'utilizzo inappropriato. ATTENZIONE: Leggere tutte le avvertenze di sicurezza prima di installare, connettere o eseguire manutenzione a questa apparecchiatura. La mancata osservanza di dette avvertenze può causare lesioni personali e/o danni alla proprietà. Conservare le istruzioni per riferimenti futuri. 1.4.3 Provvedimenti per la vostra sicurezza AVVERTENZA: Non lavorare mai in un ambiente soggetto a pericolo di esplosione! La sede dello strumento non è a tenuta di gas. (pericolo di esplosione dovuto alla formazione di scintille, corrosione provocata dall'ingresso di gas) AVVERTENZA: Quando si utilizzano prodotti chimici e solventi, conformarsi alle istruzioni del produttore e alle regole generali AVVERTENZA: Questa centrifuga non è sigillata. Utilizzare gli appropriati provvedimenti di sicurezza quando si utilizza la centrifuga per campioni infettivi e patogeni, Quando si maneggiano questi campioni, seguire le appropriate precauzioni di sicurezza. 1.4.4 Escludere le influenze ambientali seguenti       Forti vibrazioni Luce solare diretta Umidità atmosferica superiore all'80% Presenza di gas corrosivi Temperature inferiori a 2 °C e superiori a 35 °C Forti campi elettrici o magnetici: FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-3 ATTENZIONE: Pericolo di scosse elettriche all'interno della sede La sede deve sempre essere aperta solo da personale autorizzato e qualificato prima di aprire, scollegare tutte le connessioni di alimentazione dal gruppo. 1.4.5 Provvedimenti per la sicurezza durante il funzionamento              Non svitare le due metà della sede. Asciugare immediatamente eventuali versamenti di liquidi! Lo strumento non è a tenuta stagna. Verificare che la tensione in ingresso all'apparecchiatura e il tipo di spina siano compatibili con l'alimentazione locale. Connettere il cavo di alimentazione a una presa di alimentazione adeguatamente messa a terra. Utilizzare solo un cavo di alimentazione con una classificazione superiore alle specifiche dell'etichetta dell'apparecchiatura. Non posizionare l'apparecchiatura in modo che sia difficile scollegare il cavo di alimentazione dalla presa di alimentazione. Accertarsi che il cavo di alimentazione non costituisca un potenziale ostacolo o rischio di inciampo. L'apparecchiatura è per solo utilizzo all'interno. Utilizzare l'apparecchiatura solo in ambienti asciutti. Utilizzare solo accessori del tipo approvato. Azionare l'apparecchiatura solo in condizioni ambientali specificate nelle presenti istruzioni. Disconnettere l'apparecchiatura dall'alimentazione durante la pulizia. Non mettere l'apparecchiatura in funzione in ambienti pericolosi o instabili. La manutenzione deve essere esclusivamente eseguita da personale autorizzato. 1.4.6 Pericoli e precauzioni Al fine di proteggere le persone e l'ambiente, si devono osservare le precauzioni seguenti:  Durante la centrifugazione, la presenza di persone e il deposito di materiali pericolosi sono proibiti nel raggio di 30 cm dalla centrifuga conformemente alla normativa EN 61010-2-020.20.  FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R non sono a prova di esplosione e perciò non devono essere messe in funzione in zone o ambienti in cui vi sia rischio di esplosione. La centrifugazione di sostanze infiammabili, esplosive, radioattive o tali da reagire chimicamente con energia elevata è strettamente proibita. La decisione finale sui rischi associati all'utilizzo di tali sostanze è di responsabilità dell'utente della centrifuga.  Non centrifugare mai materiali tossici o patogeni senza le adeguate precauzioni di sicurezza, ovvero la centrifugazione di contenitori/tubi con chiusura emetica mancante o difettosa è strettamente proibita. L'utente ha l'obbligo di eseguire le adeguate procedure di disinfezione nel caso in cui sostanze pericolose abbiano contaminato la centrifuga e/o i suoi accessori. Quando si centrifugano sostanze infettive, prestare sempre attenzione alle precauzioni generali di laboratorio. Se necessario, mettersi in contatto con il vostro responsabile della sicurezza!  E' proibito utilizzare la centrifuga con rotori differenti da quelli elencati per questo gruppo.  Non aprire in nessun caso il coperchio della centrifuga quando il rotore ruota ancora una velocità > 2m/s 1.4.7 Abbreviazioni utilizzate nel presente manuale Simbolo/Abbreviazione giri/minuto RCF Unità -1 [min ] rpm Descrizione [x g] forza centrifuga relativa PCR giri/minuto Reazione a catena polimerasi PP - Polipropilene PC - Policarbonato accel - accelerazione decel - decelerazione prog - programma IT-4 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 2. INSTALLAZIONE 2.1 Disimballo Rimuovere attentamente la vostra centrifuga e ogni componente dall'imballo. I componenti inclusi variano secondo il modello di centrifuga (vedere tabella qui di seguito). Conservare l'imballaggio per garantire lo stoccaggio e il trasporto in condizioni di sicurezza. Il manuale di istruzioni deve sempre essere conservato con la centrifuga! Gli accessori del/dei rotore/i saranno imballati separatamente AVVERTENZA: Pericolo da sollevamento Vi è rischio di lesioni si sollevata da una sola persona Utilizzare un dispositivo di sollevamento meccanico oppure procedure di sollevamento in equipe quando si solleva o si sposta l'apparecchiatura. Fare riferimento alla sezione 8.3 per i particolari su come sollevare per estrarre dall'imballaggio. 2.1.1 Imballaggio di consegna Quantità Descrizion e 1 Centrifuga FC5714/F C5718/F C5718R/F C5816/F C5816R/FC5916/FC5916R Cavo di alimentazione Sc heda Garanzia Manuale di Is truzioni Guida Rapida Chiave Rotor e 1 1 1 1 2.2 Selezionare l'Ambiente NOTA! Evitare vibrazioni eccessive, fonti di calore, correnti d'aria o rapidi cambiamenti di temperatura.    La centrifuga deve essere installata su una superficie uniforme, salda e livellata, se possibile su un armadio/tavole da laboratorio o su una qualche altra superficie salda prova di vibrazioni. Durante la centrifugazione, la centrifuga deve essere posta di modo che vi sia uno spazio minimo di 30 cm su ogni lato del gruppo, conformemente alla normativa EN 61010-2-020. Non posizionare la centrifuga in prossimità di una finestra o di un riscaldatore, ove potrebbe essere esposta a calore eccessivo, poiché le prestazioni del gruppo si basano su una temperatura ambiente di 23°C. 2.3 Installazione Seguire le fasi seguenti:  Verificare che l'alimentazione corrisponda a quella specificata sull'etichetta dei valoro nominali del costruttore, che è posizionata sul pannello posteriore.  Per FC5714/FC5718/FC5816, la linea di alimentazione deve essere protetta da un ruttore di circuito di valore 10 A (tipo K).        Per FC5718R e FC5816R, la linea di alimentazione deve essere protetta da un ruttore di circuito di valore 16 A (tipo K). In caso di emergenza, vi deve essere un interruttore di emergenza installato all’esterno della stanza al fine di disconnettere l'alimentazione dal gruppo. Connettere la centrifuga a una presa di alimentazione messa a terra. Connettere la centrifuga alla rete (La presa per il cavo di alimentazione deve essere facilmente raggiungibile per essere disconnessa) Accendere lo strumento utilizzando l'interruttore di alimentazione di rete. Aprire il coperchio utilizzando il pulsante Door Open Rimuovere il dispositivo di fissaggio del motore durante il trasporto. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-5 2.4 Precauzioni di sicurezza durante il funzionamento e garanzia           Non azionare la centrifuga nel caso in cui non sia installata correttamente Non Sporgersi sulla centrifuga durante il funzionamento Non rimanere nel raggio di 30 cm dall''area di ingombro della macchina più a lungo del tempo necessario per ragioni operative. Non porre materiali potenzialmente pericolosi nel raggio di 30 cm dalla zona di ingombro. Non azionare la centrifuga quando smontata (es: senza sede). Non azionare la centrifuga se i componenti meccanici o elettrici sono stati manipolati. Non utilizzare accessori quali rotori e contenitori che non siano esclusivamente approvati da OHAUS Corporation, salvo tubi centrifuga di vetro o plastica disponibili in commercio. Non trattare materiali estremamente corrosivi, che possono danneggiare o indebolire i materiali. Non azionare la centrifuga con rotori o recipienti che indichino eventuali segni di corrosione o danni meccanici. Il costruttore è responsabile per la sicurezza e l'affidabilità della centrifuga solo se: 1) Il gruppo è fatto funzionare conformemente al presente manuale di istruzioni. 2) Modifiche, riparazioni o altre regolazioni sono eseguite da personale autorizzato OHAUS e l'impianto elettrico è conforme al relativo codice elettrico. NOTA! Garanzia La centrifuga è stata sottoposta ad approfondito collaudo e controllo qualità. Nell'improbabile caso di difetti di fabbricazione, la centrifuga e i rotori sono coperti da garanzia. Detta garanzia decade in caso di errata manipolazione, danni e negligenze e, inoltre nel caso di utilizzo di parti di ricambio e/o accessori inappropriati, oppure modifiche non autorizzate al gruppo. I diritti alle modifiche tecniche sono riservate al costruttore per quanto attiene miglioramenti tecnici. 3. FUNZIONAMENTO 3.1 Montaggio e caricamento del rotore 3.1.1 Installazione dei rotori Pulire l'albero motore e la bussola di chiusura con un panno pulito e privo di grasso. Posizionare il rotore sull'albero motore. (Vedere figura qui di seguito). Prestare attenzione che il rotore sia completamente installato sull'albero motore. Albero motore e camera Figura 8 Dado per Rotore Figura 9 Utensile per rotore con dado Utensile per rotore senza dado Coperchio a scatto Figura 10 Coperchio avvitabile IT-6 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Tenere il rotore con una mano e fissarlo all'albero ruotando il dado di fissaggio in senso orario. Serrare il dado di fissaggio con la chiave rotore fornita (Vedere figure 9-10) Forniremo un utensile per fissare il rotore alla centrifuga, l'utensile per il fissaggio del dado al rotore sarà fornito con il rotore ATTENZIONE! Verificare che la vite di fissaggio sia adeguatamente installata prima di ogni utilizzo. (Vedere figura 910) Non azionare la centrifuga con rotori o recipienti che indichino eventuali segni di corrosione o danni meccanici. Non operare con sostanze molto corrosive, che potrebbero danneggiare il rotore, i recipienti e i materiali. In caso di domande, si prega di prendere contatto con il costruttore! 3.1.2 Carico dei rotori d'angolo I rotori devono essere caricati simmetricamente e con lo stesso peso (Vedere figura qui di seguito). L'adattatore può essere caricato solo con recipienti approvati. La differenza di peso fra i recipienti pieni deve essere la minima possibile. Consigliamo perciò di pesarli con una bilancia. Ciò consente di ridurre l'usura della motorizzazione e il rumore durante il funzionamento. Il carico massimo per foro è dichiarato su ogni rotore. Figura 11ERRATO Figura 12 CORRETTO (6 tubi) 3.1.3 Carico dei rotori a battente Il carico dei recipienti deve essere eseguito conformemente alla figura qui di seguito. E' solo consentito operare, per esempio, con un rotore da 4 posizioni con solo 2 contenitori caricati. I contenitori caricati devono però essere opposti gli uni agli altri. Accertarsi che i contenitori scaricati siano messi all'interno del rotore (vedere qui sotto). In linea di principio, i rotori a battente non possono essere messi in funzione fino a che tutti i recipienti o le rastrelliere siano posizionati nel rotore. Le viti sul rotore devono essere lubrificate con olio "High TEF". I tubi campione devono essere riempiti uniformemente a occhio nei fori o nelle rastrelliere dei tubi. La differenza di peso dei recipienti caricati non deve essere superiore a 1,0 g circa. ATTENZIONE! I rotori a battente devono essere messi in funzione solo se tutte le sedi sono riempite con quattro recipienti o quattro attrezzi: non mischiare recipienti e attrezzi! ATTENZIONE! Non azionare la centrifuga con rotori o recipienti che indichino eventuali segni di corrosione o danni meccanici. Non operare con sostanze molto corrosive, che potrebbero danneggiare il rotore, e i recipienti. In caso di domande, si prega di prendere contatto con il costruttore! FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Figura 13 ERRATO IT-7 Figura 14 CORRETTO 3.1.4 Carico e sovraccarico dei rotori Tutti i rotori approvati sono elencati con la loro velocità max. e peso max. di riempimento in "tabella 2 peso netto ammissibile" (Vedere APPENDICE). Il carico max. ammesso per un rotore, che è determinato dal costruttore, e anche la massima velocità ammessa per questo rotore (Vedere etichetta sul rotore), non devono essere ecceduti. I liquidi con cui si carica il rotore devono avere una densità omogenea max. di 1/2 g/l o inferiore quando il rotore funziona alla velocità max. Per centrifugare liquidi con una densità più elevata, occorre ridurre la velocità conformemente alla formula seguente: Velocità ridotta nred = 1,2 √ℎ𝑖𝑔ℎ𝑒𝑟 𝑑𝑒𝑛𝑠𝑖𝑡𝑦 x velocità max. (nmax) del rotore Esempio: nred = √ 1,2 x 4.000 = 3.360 rpm 1,7 In caso di domande, si prega di prendere contatto con il costruttore! Al fine di determinare la forza centrifuga relativa (RCF/forza di accelerazione) per un adattatore specifico, è possibile calcolare secondo DIN 58 970 utilizzando la formula allegata: -5 2 RCF = 1.117862*10 *n *rmax n: giri/minuto (RPM) rmax: raggio di centrifuga max in cm utilizzando il fondo dei tubi 3.1.5 Rimozione del rotore Svitare completamente il dado di fissaggio del rotore (vite sul punto rigido) e sollevare verticalmente il rotore fuori dalla centrifuga. (Vedere figure 9 e 10) 3.2 Controllo del coperchio 3.2.1 Apertura del coperchio Dopo il funzionamento, quando il coperchio della centrifuga è chiuso, appare la parola "close" (M1) sul display "RPM | RCF" (A-1). Inoltre, se vi è un rotore nella centrifuga, appare la parola "rotor"(M3) e anche il numero di codice del rispettivo rotore che si trova all'interno del sistema centrifugo "71" (M4). Se nella centrifuga non vi è alcun rotore, la parola "rotor" (M3) lampeggia e appare la parola supplementare "no" (M4). Premendo il tasto "Door Open" (7), si può rilasciare il coperchio della centrifuga. Non appena il coperchio elettromagnetico è completamente rilasciato, appare la parola "open" (M2). Adesso è possibile aprire il coperchio della centrifuga Fare riferimento alla figura 15 qui di seguito per riferimento IT-8 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R M2 M4 M3 M1 A-1 1 7 Figura 15 Durante il funzionamento, è possibile richiamare in qualunque momento il tipo di rotore premendo il tasto "Door Open" (7). 3.2.2 Bloccaggio del coperchio Il coperchio deve solo essere abbassato leggermente Un bloccaggio elettromagnetico del coperchio fa chiudere il coperchio stesso, e allo stesso tempo scompare la parola "open" (M2) (fare riferimento alla figura 15). Quale segnale che la centrifuga è pronta per essere avviata, il display "RPM | RCF" (A-1) visualizza la parola "close" (M1) . Allo stesso tempo, si visualizza la parola "rotor" (M3), come il numero di codice del rotore che si trova nell'impianto della centrifuga, "no 71" (M4). Con questo, tutti i dati specifici del rotore, quali, per esempio. la velocità max. l'accelerazione, ecc. sono adottati. ATTENZIONE: Non lasciare impigliare le dita fra il coperchio e il dispositivo o il meccanismo di bloccaggio quando si chiude il coperchio. 3.3 Preselezione 3.3.1 Preselezione della velocità / valore RCF Questa preselezione è attivata mediante il tasto "RPM | RCF" (4) (fare riferimento alla figura 16 qui di seguito). Premendo una volta il tasto, la parola "rpm" (M5) lampeggia. Premendo due volte il tasto, è possibile selezionare la preselezione delle forze centrifughe. Poi compare la parola "rcf" (M6) lampeggiante. Si possono impostare i valori desiderati mediante la manopola di regolazione (1). Nel display (A-1), il valore regolato è visualizzato in permanenza, prima, durante e dopo M5 il funzionamento. M6 A-1 1 4 5 Figura 16 Fino a che non vi è rotore inserito, la velocità è regolabile fra 200 giri/minuto e la rotazione massima della centrifuga. Se vi è un rotore nella centrifuga, la velocità può essere preselezionata solo fino alla massima rotazione ammessa per quel rotore. Lo stesso vale con la preselezione del valore RCF. L'intervallo di impostazione è compreso tra la forza centrifuga a 200 rpm e la forza centrifuga massima consentita del rotore. Vedere "Tabella 3: velocità max. e valori RCF per il rotore ammesso " (Vedi APPENDICE). Tutti i valori importanti sono elencati lì. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-9 ATTENZIONE: Si prega anche di verificare con il costruttore il numero ammesso di rotazioni delle vostre provette. 3.3.2 Preselezione del tempo di funzionamento Il tempo di funzionamento può essere preselezionato in tre gamme differenti, da 10 secondi fino a 99 ore e 59 minuti. 1. Gamma da 10 secondi fino a 59 minuti e 50 secondi in passi di 10 secondi 2. Gamma da 1 secondi fino a 99 minuti e 59 minuti in passi di 1 minuto 3. Il funzionamento continuo "cont", può essere interrotto con il tasto "Stop"(10) (fare riferimento alla figura17). Il tempo di funzionamento può essere preselezionato con il coperchio aperto o chiuso. Per attivare le impostazioni del tempo di funzionamento, premere il tasto "Time" (6). Nel display "Time/Prog" (A-3), l'indicazione "m : s" o "h : m" lampeggia, secondo l'impostazione precedente. Per impostare il valore desiderato, utilizzare la manopola di regolazione (1). Dopo aver superato 59 minuti e 50 secondi, l'indicazione cambia automaticamente a "h : m". Dopo aver superato 99 ore 59 minuti, la parola "cont" appare nel display "Time/Prog" (A-3). Il funzionamento continuo può essere interrotto solo mediante il tasto "Stop"(10). Il conteggio alla rovescia del tempo incomincia non appena si raggiunge la velocità impostata. Il display indica sempre il tempo di funzionamento rimanente. (Vedere figura 17) A-3 1 10 6 Figura 17 3.3.3 Preselezione dell'intensità del freno e dell'accelerazione Questa funzione è attivata mediante il tasto "Accel/Decel" (5) (fare riferimento alla figura 18). Premendo il tasto una volta, la parola "accel" (M7) lampeggia nel display "Acc/Dec" (A-2). L'accelerazione desiderata può essere preselezionata mediante la manopola di regolazione (1). Il valore 0 è equivalente al più basso e il valore 9 è l'accelerazione più elevata. Premendo due volte il tasto "Accel/Decel" (5), il display "Acc/Dec" (A-2), visualizza la parola "decel"(M8). Adesso è possibile preselezionare l'intensità di frenata desiderata mediante la manopola di regolazione (1). Il valore 9 è equivalente al più breve e il valore 0 è il tempo di frenata più lungo. Vedere "tabella 5: tempi di accelerazione e decelerazione " (APPENDICE). Qui sono indicati i tempi di accelerazione e decelerazione per le fasi di accelerazione e decelerazione da 0 a 9 per i rotori ammessi. M7 M8 A-2 1 5 Figura 18 IT-10 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R /FC5916/FC5916R 3.3.4 Preselezione della temperatura (Solo Modelli Refrigerati) Questa funzione è attivata dal tasto "Temp/Setup" (13). Dopo aver premuto questo tasto, nel display "Time/Prog" lampeggia l'indicazione "°C" (A-4). Mediante la manopola di regolazione (1) è possibile preselezionare la temperatura di test in passi da 1°C in una gamma da -20°C fino a +40°C. Questo valore è indicato in modo permanente nel display (figura 19), prima durante e dopo il funzionamento. Si prega di notare le rispettive temperature più basse dei rotori alla velocità max.! A-4 1 13 Figura 19 3.3.5 Pre-raffreddamento (Solo Modelli Refrigerati) Se i campioni sono sensibili alla temperatura, è utile preraffreddare la centrifuga, il rotore e in ultimo i contenitori alla temperatura di funzionamento necessaria. Perciò, inserire il rotore desiderato e preimpostare la rispettiva temperatura. Si avvia il ciclo premendo simultaneamente i tasti "Temp/Setup" (13) (fare riferimento alla figura 20) e "Time" (6). Durante il funzionamento, il gruppo sceglie automaticamente una velocità di rotazione equivalente al 20% della velocità di rotazione del rispettivo rotore. Dopo aver raggiunto la temperatura preimpostata, è possibile lasciare il ciclo di preraffreddamento con il tasto "Stop" (10). Secondo il rotore inserito, il preraffreddamento va da circa 10 a 20 minuti. A-4 13 10 6 Figura 20 3.4 Correzione del raggio Se si utilizzano adattatori o riduttori, è possibile modificare il raggio centrifugo del rispettivo rotore. Nel qual caso, il raggio può essere corretto manualmente. Procedere come segue: Chiudere il coperchio, quindi premere allo stesso tempo il tasto "Time" (6) (fare riferimento alla figura 21) e il tasto "Prog/Setup" (11) e tenerli premuti. Il display "Time/Prog" (A-3) visualizza la parola "radius" (M9). Regolando la manopola (1), è possibile preselezionare la rispettiva correzione del raggio (Vedere tabella 7, APPENDICE) in passi di 0.1 cm. Non appena si è selezionata una correzione del raggio, compare la parola "radius" (M9). Questa parola rimarrà visibile fino a che non si rimette la correzione del raggio a 0. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-11 A-3 M9 11 1 6 Figura 21 3.5 Programma 3.5.1 Salvataggio di programmi Si possono salvare fino a 99 cicli con tutti i relativi parametri, compresi i rotori utilizzati. Si può utilizzare qualunque numero di programma libero e richiamarlo. Porre nella centrifuga il rotore necessario. Premendo il tasto "Prog/Setup" (11) sul display "Time/Prog" (A-3), appare la parola "programm" . E' possibile scegliere il numero di programma desiderato mediante la manopola di regolazione (1). Se un numero di programma è già occupato, sul display "RPM | RCF" (A-1), appariranno le parole "rotor" (M3) a "xx" (M4). Nel caso di numeri di programma liberi, appare 0 M4 M3 A-3 A-1 M10 11 1 12 Figura 22 Chiudere il coperchio della centrifuga Adesso procedere come descritto in precedenza per impostare tutti i parametri di ciclo importanti. Se il coperchio non è chiuso quando si salva il programma, sul display "RPM/RCF" (A-1), lampeggiano alternativamente le parole "FirSt" e "CLOSE Lid" (Vedere figura 23). Se si vuole avviare il ciclo senza salvare il programma, sul display "RPM/RCF" (A-1) lampeggiano alternativamente le parole "First" e "PrESS StoreE" (Vedere figura 24) . Figura 23 Figura 24 IT-12 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Per adottare i dati, premere il tasto "Store" (12) (fare riferimento alle figure 23 e 24) per circa 1 secondo. Se il programma è salvato correttamente, la parola "Store" appare sul display "RPM/RCF" (A-1). Per effetto di questo, la parola "programm" (M10) scompare. Non appena si rilascia il tasto "Store" (12), ricompare la parola "programm xx" (M10) , ove (xx) sta per la posizione scelta per il programma. Se tutti i numeri di programma sono occupati, si può prendere un vecchio numero che non sia più necessario e semplicemente digitare i nuovi parametri. 3.5.2 Richiamo dei programmi salvati Pre richiamare i programmi salvati, premere il tasto "Prog/Setup" (11) (fare riferimento alla figura 25) quando il coperchio è già chiuso. Sul display "Time/Prog" (A-3), appare la parola "programm –"(M10). Il numero di programma desiderato può essere preselezionato mediante la manopola di regolazione (1). Nei rispettivi display appariranno i valori salvati per quel programma. Se all'interno della centrifuga si trova un rotore errato per il programma preselezionato, sul display "RPM | RCF" (A-1), lampeggerà la parola "rotor" (M3) . Nello stesso tempo, lampeggiano a turno la parola "FALSE" e il numero di rotore salvato " xx" (M4). M3 M4 A-3 A-1 M10 11 1 12 Figura 25 3.5.3 Lasciare la modalità programma Per lasciare la modalità programma, premere semplicemente il tasto "Prog/Setup" (11) (fare riferimento alla fig. 25). Sul display "Time/Prog" appare la parola "programm" . Impostare il display a "programm--" (M10) con la manopola di regolazione (1). 3.6 Avviamento e arresto della centrifuga 3.6.1 Avviare la centrifuga E' possibile avviare la centrifuga con il tasto "Start" (9) (rif. figura 26) oppure con il tasto "Quick Spin" (8). Con il tasto "Start" (9), si possono avviare cicli salvati oppure cicli con i parametri preselezionati manualmente. Quando è finito il rispettivo tempo ciclo preselezionato, la centrifuga si arresterà automaticamente. Con il tasto "Quick Spin" (8), si possono avviare cicli, che dureranno appena qualche secondo. Premendo il tasto "Quick Spin" (8), la centrifuga accelera fino alla rotazione selezionata. Sul display "Time/Prog" (A-3), il tempo ciclo trascorso è indicato a partire da quando si è premuto il tasto "Quick Spin" (8). Rilasciando il tasto "Quick Spin" (8), la centrifuga si arresta e il tempo ciclo è indicato fino all'apertura del coperchio. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-13 A-3 9 1 8 Figura 26 3.6.2 Il tasto "STOP" Mediante il tasto "Stop" (10) (Vedere figura27) è possibile interrompere il ciclo in ogni momento. Dopo aver premuto il tasto, la centrifuga decelera all'intensità preselezionata fino a che non si arresta. 10 Figura 27 3.7 Rilevamento di squilibri Nel caso in cui il rotore non sia caricato uniformemente, la motorizzazione si disconnetterà durante la decelerazione. Il rotore decelera fino a che non si arresta. Quando sul display "Time/Prog" (A-3) appare la parola "error” (M11) insieme al numero "01” , la differenza dei campioni è eccessiva. Distribuire uniformemente il peso. Caricare il rotore come descritto ai capitoli 3.1.2 e 3.1.3. Quando sul display "Time/Prog" (A-3) appare la parola "error” insieme al numero "02" (Vedere figura 28), ciò può essere dovuto alla ragione seguente. Difetto dell'interruttore di squilibrio. A-3 M11 Figura 28 IT-14 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4. IMPOSTAZIONE 4.1 Regolazioni di base 4.1.1 Accesso alla modalità "Dati di funzionamento" Quando si utilizza la centrifuga, è possibile impostare i parametri seguenti: - Indicazione di temperatura °C o °F - Accensione/spegnimento del segnale acustico - Accensione/spegnimento del suono tastiera - Pre-selezione del volume del segnale acustico - Selezione della suoneria del segnale acustico "fine ciclo" Mentre la centrifuga è disconnessa, premere simultaneamente i tasti "Time"(6) e "Door Open" (7) e accendere l'interruttore principale della centrifuga. Adesso, rilasciare entrambi i tasti e - per effetto di ciò - viene eseguita una prova del display per circa 5 secondi. Tutti gli indicatori appariranno allo stesso tempo (Vedere figura 29). 7 6 Figura 29 ATTENZIONE: Si noti che è necessario inserire il programma descritto al punto 4.1.1 per modificare le regolazioni dei punti 4.1.2 - 4.1.7. Dopo che si sono salvate le impostazioni, ritornare nuovamente alla modalità normale disconnettendo la centrifuga per un breve tempo. Tutte le impostazioni modificate devono essere confermate con il tasto "Start"(9). La parola "Store"(12) appare sul display "RPM | RCF"(A-1) - Nelle preselezioni sono validi solo! A-1 9 12 Figura 30 4.1.2 Indicazione della temperatura Procedere come descritto al punto 4.1.1 per inserire questa modalità programma e poi premere il tasto "Accel/Decel" (5). Sul display "Acc/Dec" (A-2) compare la parola "Service" . Selezionare adesso la lettera "C" con la manopola di regolazione (1). Di conseguenza, sul display "RPM | RCF" (A-1), appaiono le parole "CELSI/temp" . Se si preme il tasto "RPM | RCF" (4), lampeggia la parola "CELSI" e si può modificare il display Fahrenheit "FAREN", con la manopola di regolazione(1) (Vedere figura 31). Dopo che si sono salvate le impostazioni (Vedere 4.1.1) ritornare nuovamente alla modalità normale disconnettendo la centrifuga per un breve tempo. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-15 1 4 5 Figura 31 4.1.3 Accensione/Spegnimento segnale Procedere come descritto al punto 4.1.1 per inserire questa modalità programma e poi premere il tasto "Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" . Selezionare adesso la lettera "L con la manopola di regolazione (1). Di conseguenza, compaiono le parole "On Sound" sul display "RPM | RCF" (4). Se adesso si preme il tasto "RPM | RCF" (4), lampeggia la parola "On" e si può disattivare la suoneria, con la manopola di regolazione(1) (Vedere figura 32). Dopo che si sono salvate le impostazioni (Vedere 4.1.1) ritornare nuovamente alla modalità normale disconnettendo la centrifuga per un breve tempo. A-1 A-2 1 4 5 Figura 32 4.1.4 Pre-selezione del volume del segnale acustico Procedere come descritto al punto 4.1.per inserire questa modalità programma e poi premere il tasto "Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" . Selezionare adesso la lettera "U con la manopola di regolazione (1). Di conseguenza, sul display "RPM | RCF" (A-1), appaiono le parole "Vol=09/Sound" . Dopo aver premuto il tasto "RPM | RCF" (4), è possibile regolare il volume desiderato fra 0 (basso) e 9 (alto) con la manopola di regolazione (1) (Vedere figura 33). Dopo che si sono salvate le impostazioni (Vedere 4.1.1) ritornare nuovamente alla modalità normale disconnettendo la centrifuga per un breve tempo. A-1 A-2 1 4 5 Figura 33 IT-16 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 4.1.5 Selezione della suoneria del segnale acustico "fine ciclo" Procedere come descritto al punto 4.1.per inserire questa modalità programma e poi premere il tasto "Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" . Selezionare adesso la lettera "G con la manopola di regolazione (1). Di conseguenza, sul display "RPM | RCF" (A-1), appare la parola "SonGo/Sound" . Dopo aver premuto il tasto "RPM | RCF" (4), è possibile selezionare una suoneria con la manopola di regolazione(1). (Vedere figura 34) Dopo che si sono salvate le impostazioni (Vedere 4.1.1), si può ritornare nuovamente alla modalità normale disconnettendo la centrifuga per un breve tempo. A-1 A-2 1 4 5 Figura 34 4.1.6 Accensione/spegnimento del suono tastiera Procedere come descritto al punto 4.1.per inserire questa modalità programma e poi premere il tasto "Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" . Selezionare adesso la lettera "B con la manopola di regolazione (1). Di conseguenza, sul display "RPM | RCF" (A-1), appare la parola “ON/BEEP . Dopo aver premuto il tasto "RPM | RCF" (4), è possibile selezionare il suono tastiera con la manopola di regolazione(1). (Vedere figura 39). Dopo che si sono salvate le impostazioni (Vedere 4.1.1), si può ritornare nuovamente alla modalità normale disconnettendo la centrifuga per un breve tempo. A-1 A-2 1 4 5 Figura 35 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-17 4.1.7 Richiamo dei dati di funzionamento ATTENZIONE: Questa operazione deve essere eseguita solo da utenti avanzati o da un tecnico di manutenzione. I dati di funzionamento della centrifuga possono essere richiamati nella modalità "Basic Adjustments" . Per inserire questa modalità di programma, procedere come descritto al punto 4.1.2. Premere il tasto "Accel/Decel" (5). Sul display "Acc/Dec" (A-2) lampeggia la parola "Service" . E' possibile avere accesso alle differenti informazioni mediante la manopola di regolazione (1): A= avviamenti precedenti della centrifuga H= ore precedenti di funzionamento S= versione software r= software di conversione E= elenco dei messaggi di errore precedenti h= tempo di funzionamento del motore E' possibile visualizzare l'elenco degli ultimi 99 messaggi di errore premendo il tasto "RPM | RCF" (4) e scorrerlo con la manopola di regolazione (1). I rispettivi codici di errore compaiono nel display "RPM | RCF" (A-1). Fare riferimento a "Tabella 6: messaggio di errore " (Vedere APPENDICE). Per ritornare ala modalità programma normale, spegnere la centrifuga per un breve tempo. A-1 A-2 1 4 5 Figura 36 5. MANUTENZIONE 5.1 Manutenzione e pulizia 5.1.1 Generalità Attenzione: La manutenzione della centrifuga si limita al tener puliti il rotore, la camera del rotore e gli accessori del rotore stesso, e anche a lubrificare regolarmente le viti di inserto del rotore per un rotore a battente (se disponibile). Il lubrificante più appropriato è l'olio High TEF. Lubrificanti contenenti molycote e grafite non sono consentiti Si prega di prestare particolare attenzione alle parti in alluminio anodizzato. Anche piccoli danni possono provocare la rottura del rotore. Nel caso in cui rotori, contenitori o rastrelliere di tubi vengano in contatto con sostanze corrosive, i punti interessati devono essere puliti con attenzione. Sostanze corrosive sono, per esempio: alcali, soluzioni di sapone, amine alcaline, acidi concentrati, soluzioni contenenti metalli pesanti, solventi clorati privi di acqua, soluzioni saline, quali acqua salata, fenolo e idrocarburi alogenati. IT-18 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Pulizia – gruppi, rotori e accessori: - Spegnere il dispositivo e disconnetterlo dall'alimentazione prima di incominciare qualunque operazione di pulizia e disinfezione. Non versare liquidi all'interno della sede Non aspergere disinfettante sul dispositivo. Un'attenta pulizia non serve solo a disinfettare, ma anche a evitare la corrosione dovuta all'inquinamento. Al fine di evitare di danneggiare parti anodizzate quali rotori, piastre di riduzione, ecc. per la pulizia è possibile utilizzare solo detergenti a pH neutro con un valore pH- da 6 a 8. Non si devono utilizzare agenti detergenti alcalini (valore pH > 8). Dopo la pulizia, accertarsi che tutti gli elementi siano ben asciutti, manualmente o mediante una cassetta ad aria riscaldata (Temperatura max.+ 50°C). E' necessario rivestire regolarmente gli elementi in alluminio anodizzato con olio anti-corrosione al fine di aumentare la loro durata in vita e di ridurre la predisposizione alla corrosione. Per via dell'umidità o di campioni non ermeticamente chiusi, è possibile si formi della condensa. La condensa deve essere rimossa regolarmente dalla camera del rotore con un panno soffice. La procedura di manutenzione deve essere ripetuta ogni 10-15 funzionamenti, o almeno una volta la settimana. - Connettere il gruppo all'alimentazione, una volta che il macchinario sia completamente asciutto. Non eseguire disinfezioni con raggi UV, beta e gamma o con altre radiazioni a energia elevata. I rotori di metallo possono essere puliti in autoclave . Anche il coperchio e gli adattatori del rotore possono essere puliti in autoclave (max. 121°C, 20 min). Le rastrelliere dei tubi sono in polipropilene e non possono essere puliti e in autoclave a 134°C. 5.1.2 Pulizia e disinfezione del gruppo 1. Aprire il coperchio prima di spegnere il gruppo. Disconnetterlo dall'alimentazione 2. Aprire il dado del rotore ruotando in senso antiorario la chiave del rotore. 3. Rimuovere il rotore. 4. Per pulire e disinfettare il gruppo e la camera del rotore, utilizzare l'agente di pulizia citato sopra. 5. Pulire tutte le zone accessibili del dispositivo e dei suoi accessori, compreso il cavo di alimentazione, con un panno umido. 6. Lavare a fondo le guarnizioni di gomma e la camera del rotore con acqua. 7. Cospargere le guarnizioni di gomma asciutte con glicelolo o talco al fine di prevenire che divengano fragili. Altri componenti del gruppo, per esempio il bloccaggio del coperchio, l'albero motore e il rotore, non devono essere lubrificati. 8. Asciugare l'albero motore con un panno morbido, asciutto e senza sfilacciature. 9. Verificare che il gruppo e gli accessori non abbiano subito danni. Accertarsi che la centrifuga sia spenta e disconnettere il gruppo dall'alimentazione. Rimuovere quindi la polvere che aderisce alle asole di ventilazione nella centrifuga mediante una spazzola soffice. Eseguire questa operazione almeno ogni sei mesi. 5.1.3 Pulizia e disinfezione del rotore 1. Pulire e disinfettare i rotori, i coperchi del rotore e gli adattatori con l'agente di pulizia citato sopra. 2. Utilizzare uno scovolo per pulire e disinfettare gli alesaggi del rotore. 3. Risciacquare i rotori, il coperchio del rotore e l'adattatore con acqua pulita. In particolare, i fori dei rotori ad angolo. 4. Per asciugare i rotori e gli accessori, porli su un telo. Posizionare i rotori ad angolo con gli alesaggi verso il basso 5. Asciugare il cono del rotore con un panno morbido, asciutto e senza sfilacciature e verificare la presenza di eventuali danni. Non lubrificare il cono del rotore. 5.1.4 Disinfezione dei rotori in alluminio Nel caso in cui materiale infettivo penetri nella centrifuga, il rotore e la camera del rotore devono essere disinfettati immediatamente dopo il funzionamento. I rotori possono essere puliti in autoclave alla temperatura max. di 121° C. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-19 5.1.5 Disinfezione dei rotori in polipropilene Pulizia in autoclave Il tempo consigliato per la pulizia in autoclave da 15 a 20 min a 121°C (1 bar) ATTENZIONE: Il tempo di sterilizzazione di 20 minuti non deve essere ecceduto. Una sterilizzazione ripetuta provocherà una riduzione della resistenza meccanica della materia plastica. Prima di essere passato in autoclave, il rotore e l'adattatore in polipropilene deve essere pulito attentamente al fine di evitare che i residui di sporcizia brucino. Le conseguenze di alcuni residui di prodotti chimici nei materiali plastici a temperatura ambiente possono essere trascurate. A temperature elevate, però, durante il passaggio in autoclave questi residui possono corrodere e distruggere la plastica. Gli oggetti devono essere risciacquati a fondo con acqua distillata dopo la pulizia, ma prima del passaggio in autoclave, I residui di qualunque liquido di pulizia possono provocare fessurazioni, sbiancamento e macchie. Sterilizzazione a gas Gli adattatori, le bottiglie e i rotori possono essere sterilizzati con gas di ossido di Etilene Accertarsi di aerare gli articoli dopo la sterilizzazione e prima di utilizzarli nuovamente. ATTENZIONE: Poiché durante la sterilizzazione la temperatura può salire, i rotori, gli adattatori e le bottiglie non devono essere chiuse e devono essere completamente svitate. Sterilizzazione chimica Le bottiglie, gli adattatori e i rotori possono essere trattati con i normali disinfettanti chimici. ATTENZIONE: Prima di applicare qualunque altro metodo di pulizia o di decontaminazione non espressamente consigliato dal costruttore, prendere contatto con il costruttore per garantire che non danneggi il gruppo o il rotore. 5.1.6 Rottura del vetro Con coefficienti G elevati aumenta la velocità di rottura dei tubi di vetro . I frammenti di vetro devono essere rimossi immediatamente dal rotore, dai contenitori, dagli adattatori e dalla camera stessa del rotore. Piccoli frammenti di vetro graffieranno e perciò danneggeranno il rivestimento protettivo di superficie di un rotore. Se frammenti di vetro rimangono nella camera del rotore, si accumulerà polvere metallica fine per via della circolazione dell'aria. Questa polvere metallica nera molto sottile inquinerà molto gravemente la camera del rotore, il rotore, i contenitori e i campioni. Se necessario, sostituire gli adattatori, i tubi e gli accessori al fine di evitare ulteriori danni. Verificare regolarmente che gli alesaggi del rotore non contengano residui o presentino danni. ATTENZIONE: Verificare le specifiche del costruttore dei tubi per la centrifuga. 5.2 Durata in vita di rotori, contenitori e accessori I rotori e il coperchio del rotore in alluminio o in acciaio inox hanno una durata di funzionamento max. di 7 anni dal primo utilizzo. I coperchi trasparenti del rotore in PC o PP, come anche i rotori, le rastrelliere dei tubi e gli adattatori in PP hanno una durata in vita max. di 3 anni dal primo utilizzo. Condizione per la durata in vita: Utilizzo appropriato, condizione priva di danni e cura consigliata. IT-20 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 6. RICERCA GUASTI 6.1 Messaggio di errore: Causa/Soluzione I messaggi di errore sono elencati per consentire una più rapida localizzazione degli errori. E' possibile che la diagnosi cui si fa riferimento in questo capitolo non siano sempre quelle reali, poiché si tratta di errori e soluzioni che avvengono solo teoricamente. Vi preghiamo di tenerci informati di ogni tipo di errore che si verifichi e non sia fra quelli elencati nel presente capitolo. Solo le vostre informazioni ci consentono di migliorare questo manuale di funzionamento. Vi ringraziamo anticipatamente per il vostro sostegno. 6.2 Esame dei possibili messaggi di errore e delle relative soluzioni 6.2.1 Rilascio del coperchio durante una caduta di tensione (Rilascio del Coperchio di Emergenza) In caso di caduta di tensione o di malfunzionamento, il coperchio della centrifuga può essere aperto al fine di prelevare i vostri campioni. Per FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R (bloccaggio motorizzato) Procedere come segue: 1. Spegnere la centrifuga e disconnettere il cavo di alimentazione, attendere fino a che il rotore non sia fermo (possono essere necessari alcuni minuti) 2. Sul lato sinistro della sede della centrifuga vi è un arresto di plastica Rimuovere l'arresto e dietro di esso vi è un dado esagonale. 3. Utilizzare la chiave di fornitura per la cassetta, porla nel foro e bloccare il dado esagonale con la chiave (Vedere figura 37). 4. Adesso ruotare la chiave sul lato destro (senso orario) fino al limite. ATTENZIONE: a) Ruotare semplicemente fino al limite, non serrare il dado. b) Adesso aprire il coperchio della centrifuga c) Accendere nuovamente la centrifuga, per riprendere il lavoro. Per 5714 Procedere nel modo seguente (vedere Figura 38): Figura 37 ATTENZIONE: • Spegnere la centrifuga e disconnettere il cavo di alimentazione, attendere fino a che il rotore non si arresta (possono essere necessari alcuni minuti) Sul lato destro della sede della centrifuga vi è un arresto di plastica (vedere Figura 38) Rimuovere questo arresto, che è connesso al bloccaggio del coperchio, orizzontalmente dalla sede fino a che il coperchio della centrifuga non si apre. • Adesso aprire il coperchio della centrifuga FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-21 Figura 38 6.2.2 Descrizione del sistema dei messaggi di errore Il messaggio di errore "error" (M11) è visualizzato nel display "Time/Prog" (A-3) (Vedere figura 39). Informazioni dettagliate su possibili messaggi di errore si trovano su "tabella 6: messaggi di errore " (Vedere Appendice). A-3 M11 Figura 39 7. RICEZIONE DELLE CENTRIFUGHE PER RIPARAZIONE Rischi per la salute da apparecchiature, rotori e accessori contaminati. Nel caso di ritorno al costruttore della centrifuga per riparazioni, si prega di notare quanto segue: La centrifuga deve essere decontaminata e pulita prima della spedizione per la protezione di persone, ambiente e materiale. Certificazione di decontaminazione al momento della consegna dei beni da restituire (Vedere APPENDICE) Ci riserviamo il diritto di non accettare centrifughe contaminate. Inoltre, tutti i costi sostenuti per la pulizia e la disinfezione dei gruppi saranno addebitati al cliente. IT-22 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 8. TRASPORTO, STOCCAGGIO E SMALTIMENTO 8.1 Trasporto Smontare il rotore prima del trasporto. Trasportare il gruppo solo nell'imballaggio originale. Quando si trasporta per lunghe distanze, installare il materiale da protezione del trasporto per fissare l'albero motore. General transportation Air temperature rel. humidity Air pressure -25 to 60 °C 10 to 75 % 30 to 106 kPa 8.2 Stoccaggio Durante lo stoccaggio della centrifuga, occorre rispettare le condizioni ambientali seguenti: in transport packaging Air temperature rel. Humidity Air pressure -25 to 55 °C 10 to 75 % 70 to 106 kPa 8.3 Trasporto, installazione e Smaltimento della Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R Queste istruzioni complimentano le istruzioni generali precedenti di cui al capitolo 8 e non le sostituiscono. 8.3.1 Trasporto • • Trasportare il gruppo solo nell'imballaggio originale. La centrifuga deve essere sempre trasportata utilizzando un dispositivo di trasporto meccanico. 8.3.2 Installazione  1. 2. 3. 4. Aprire il cartone e sollevare il dispositivo Tagliare il nastro adesivo Aprire le 4 alette del cartone Rimuovere gli accessori Sollevare con cura la centrifuga dal cartone Avvertenza: Pericolo da sollevamento Vi è rischio di lesioni si sollevata da una sola persona Utilizzare un dispositivo di sollevamento meccanico oppure procedure di sollevamento in equipe quando si solleva o si sposta l'apparecchiatura.  1. 2. 3. 4. Posizionare il dispositivo su un banco da laboratorio stabile, orizzontale e senza risonanze. Rimuovere il materiale di protezione anteriore e posteriore Rimuovere il manicotto di plastica. Mantenere una distanza minima di 30 cm dai dispositivi in prossimità, ai lati e dal lato posteriore alla parete. Installare il dispositivo in un luogo ben ventilato, protetto dalla luce solare diretta, al fine di impedirne il surriscaldamento. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R  Connettere il dispositivo. 1. Dopo l'installazione, attendere 4 ore prima di accendere la centrifuga, al fine di evitare danni al compressore. 2. Verificare che la tensione e la frequenza di rete corrispondano ai requisiti di targa del dispositivo (Vedere lato posteriore del dispositivo) e quindi connettere il dispositivo all'alimentazione.  Rimuovere il materiale di protezione durante il trasporto dalla camera del rotore. 1. Accendere il dispositivo sull'interruttore di alimentazione di rete. 2. Aprire il coperchio della centrifuga utilizzando il pulsante Open 3. Rimuovere il materiale di protezione di trasporto 4. Posizionare il rotore verticalmente sull'albero motore. 5. Ruotare il dado del rotore ruotando la chiave del rotore in senso orario fino a che il dado del rotore non è serrato.  Adesso il dispositivo é pronto per essere utilizzato. Conservare l'imballaggio e tutto il materiale di protezione durante il trasporto per spedire il dispositivo a una data successiva. 8.3.3 Imballaggio Imballare la centrifuga nell'ordine inverso 8.3.4 Trasferimento del dispositivo Quando si trasferisce il dispositivo a terzi, accertarsi che comprenda anche il presente manuale di istruzioni. IT-23 IT-24 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9. DATI TECNICI 9.1 Specifiche 9.1.1 Centrifuga FC5714 Modello Gamma velocità: FC5714 200 giri/minuto -14000 giri/minuto;10 giri/minuto/impostazione RCF massimo 18624 x g;10 x g/impostazione Capacità Max. (Rotore) 4x100ml Gamma Temperature (N/A) Aria fresca Tempo di funzionamento Livello di rumorosità (dipendente dal rotore) Densità consentita a max. velocità da 10 sec a 99 ore 59 min 59 sec o continuo ≤ 63 ± 2 dB(A) 1.2 g/ml Energia cinetica consentita 5595 Nm Connessione di alimentazione di rete AC Fluttuazione di tensione 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz ±10 % ±10 % Consumo di corrente 1,3 A 2,4 A Consumo di potenza 240 W 300 W 362 x 493 x 330 mm Dimensioni (W × D × H) 14,3 x 19,4 x 13,0 in 30 kg Peso Netto (senza rotore) 66 lb 580 x 490 x 460 mm Dimensioni di Spedizione (W × D × H) 22,8 x 19,3 x 18,1 in 32,5 kg Peso di Spedizione (senza rotore) 72 lb Condizioni ambiente (EN/IEC 61010-1) Ambiente Solo per uso al coperto Altitudine Utilizzabile a un'altitudine fino a 2000 m Temperatura ambiente Umidità relativa max. da 2°C a 35 °C Umidità relativa max. 80 % per temperature fino a 31°C, in diminuzione lineare al 50 % di umidità relativa fino a 35°C. Categoria sovratensione (IEC 60364-4-443) II Grado di contaminazione 2 Classe di protezione I Non adatta per l'utilizzo in ambienti pericolosi FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-25 9.1.2 Centrifuga FC571 Modello Gamma velocità: FC5718 200 giri/minuto -18000 giri/minuto;10 giri/minuto/impostazione RCF massimo 23542 x g;10 x g/impostazione Capacità Max. (Rotore) 4x100ml Gamma Temperature (N/A) Tempo di Funzionamento Livello di rumorosità Livello di rumorosità (dipendente dal rotore) Densità consentita a max. velocità Aria fresca da 10 sec a 99 ore 59 min 59 sec o continuo ≤ 60 ± 2 dB(A) 1,2 g/ml Energia cinetica consentita 16672 Nm Connessione di alimentazione di rete AC 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz Fluttuazione di tensione ±10 % ±10 % Consumo di corrente 2,0 A 4,0 A Consumo di potenza 455 W 475 W 408 x 499 x 351 mm 16,1 x 19,7 x 13,8 in 43 kg Dimensioni (W × D × H) Peso Netto (senza rotore) 95 lb 650 x 520 x 490 mm Dimensioni di Spedizione (W × D × H) 25,6 x 20,5 x 19,3 in 53 kg Peso di Spedizione (senza rotore) 117 lb Condizioni ambiente (EN/IEC 61010-1) Ambiente solo per uso al coperto Altitudine Utilizzabile a un'altitudine fino a 2000 m Temperatura ambiente Umidità relativa max. Categoria sovratensione (IEC 60364-4443) Grado di contaminazione da 2°C a 35 °C Umidità relativa max. 80 % per temperature fino a 31°C, in diminuzione lineare al 50 % di umidità relativa fino a 35°C. Classe di protezione Non adatta per l'utilizzo in ambienti pericolosi II 2 I IT-26 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.3 Centrifuga FC5718R Modello Gamma velocità: FC5718R 200 giri/minuto -18000 giri/minuto;10 giri/minuto/impostazione RCF massimo 23542 x g;10 x g/impostazione Capacità Max. (Rotore) 4x100ml Gamma Temperature (Digitale) -20~40℃ Tempo di funzionamento Livello di rumorosità (dipendente dal rotore) Densità consentita a max. velocità da 10 sec a 99 ore 59 min 59 sec o continuo ≤ 60 ± 2 dB(A) 1,2 g/ml Energia cinetica consentita 25111 Nm Connessione di alimentazione di rete AC 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz Fluttuazione di tensione ±10 % ±10 % Consumo di corrente 3,0 A 6,0 A Consumo di potenza 660 W 660 W 407 x 731 x 359 mm Dimensioni (W × D × H) 16,0 x 28,8 x 14,1 in 60 kg Peso Netto (senza rotore) 132 lb 840 x 640 x 590 mm Dimensioni di Spedizione (W × D × H) 33,1 x 25,2 x 23,2 in 77 kg Peso di Spedizione (senza rotore) 170 lb Condizioni ambiente (EN/IEC 61010-1) Ambiente solo per uso al coperto Altitudine Utilizzabile a un'altitudine fino a 2000 m Temperatura ambiente Umidità relativa max. da 2°C a 35 °C Umidità relativa max. 80 % per temperature fino a 31°C, in diminuzione lineare al 50 % di umidità relativa fino a 35°C. Categoria sovratensione (IEC 60364-4443) Grado di contaminazione Classe di protezione Non adatta per l'utilizzo in ambienti pericolosi II 2 I FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-27 9.1.4 Centrifuga FC5816 Modello Gamma velocità: FC5816 200 giri/minuto -15000 giri/minuto;10 giri/minuto/impostazione RCF massimo 21379 x g;10 x g/impostazione Capacità Max. (Rotore) 6 x 250ml Gamma Temperature (N/A) Aria fresca Tempo di funzionamento Livello di rumorosità (dipendente dal rotore) Densità consentita a max. velocità da 10 sec a 99 ore 59 min 59 sec o continuo ≤ 61 ± 2 dB(A) 1,2 g/ml Energia cinetica consentita 34363 Nm Connessione di alimentazione di rete AC 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz Fluttuazione di tensione ±10 % ±10 % Consumo di corrente 2,4 A 4,2 A Consumo di potenza 530 W 520 W 446 x 538 x 354 mm Dimensioni (W × D × H) 17,6 x 21,2 x 13,9 in 52 kg Peso Netto (senza rotore) 115 lb 840 x 640 x 590 mm Dimensioni di Spedizione (W × D × H) 33,1 x 25,2 x 23,2 in 77 kg Peso di Spedizione (senza rotore) 170 lb Condizioni ambiente (EN/IEC 61010-1) Ambiente solo per uso al coperto Altitudine Utilizzabile a un'altitudine fino a 2000 m Temperatura ambiente Umidità relativa max. da 2°C a 35 °C Umidità relativa max. 80 % per temperature fino a 31°C, in diminuzione lineare al 50 % di umidità relativa fino a 35°C. Categoria sovratensione (IEC 60364-4443) Grado di contaminazione Classe di protezione Non adatta per l'utilizzo in ambienti pericolosi II 2 I IT-28 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.5 Centrifuga FC5816R Modello Gamma velocità: FC5816R 200 giri/minuto -16000 giri/minuto;10 giri/minuto/impostazione RCF massimo 24325 x g;10 x g/impostazione Capacità Max. (Rotore) 6 x 250ml Gamma Temperature (Digitale) -20~40℃ Tempo di funzionamento Livello di rumorosità (dipendente dal rotore) Densità consentita a max. velocità da 10 sec a 99 ore 59 min 59 sec o continuo ≤ 63 ± 2 dB(A) 1,2 g/ml Energia cinetica consentita 34363 Nm Connessione di alimentazione di rete AC 230 V ~ 50/60 Hz 120 V ~ 50/60 Hz Fluttuazione di tensione ±10 % ±10 % Consumo di corrente 3,7 A 7,8 A Consumo di potenza 785 W 850 W 723 x 538 x 354 mm Dimensioni (W × D × H) 28,5 x 21,2 x 13,9 in 77 kg Peso Netto (senza rotore) 170 lb 840 x 640 x 590 mm Dimensioni di Spedizione (W × D × H) 33,1 x 25,2 x 23,2 in 87 kg Peso di Spedizione (senza rotore) 192 lb Condizioni ambiente (EN/IEC 61010-1) Ambiente solo per uso al coperto Altitudine Utilizzabile a un'altitudine fino a 2000 m Temperatura ambiente Umidità relativa max. da 2°C a 35 °C Umidità relativa max. 80 % per temperature fino a 31°C, in diminuzione lineare al 50 % di umidità relativa fino a 35°C. Categoria sovratensione (IEC 60364-4443) Grado di contaminazione Classe di protezione Non adatta per l'utilizzo in ambienti pericolosi II 2 I FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.6 Centrifuge FC5916 Modello Intervallo velocità RCF massimo Capacità massima (rotore) Intervallo di temperatura (N/A) Tempo di esecuzione FC5916 200 rpm -16000 rpm;10 rpm/impostato 24325 x g;10 x g/impostato 4X750 ml Raffreddamento ad aria 10 sec a 99 ore 59 min 59 sec o continuo Livello di rumore (a seconda del rotore) Densità consentita alla massima velocità Energia cinetica consentita Collegamento alla rete elettrica CA ≤ 63 ± 2 dB(A) 1,2 g/ml Fluttuazione di tensione Consumo di corrente Consumo di energia Dimensioni (L × P × A) Peso netto (senza rotore) Dimensioni di spedizione (L × P × A) Peso di spedizione (senza rotore) Condizioni ambientali (EN/IEC 61010-1) Ambiente Altitudine Temperatura ambiente Umidità relativa massima IT-29 60629 Nm 230 V ~ 120 V ~ 50/60 Hz 50/60 Hz ± 10 % ± 10 % 2,8 A 5,6 A 640 W 680 W 540 x 670 x 390 mm 21,3 x 26,4 x 15,4 pollici 85 kg 187 libbre 780 x 670 x 590 mm 30,7 x 26,4 x 23,3 pollici 98 kg 216 libbre Solo per uso interno Utilizzare fino ad un'altitudine di 2000 m 2°C fino a 35°C Umidità relativa max. 80% per temperature fino a 31°C, diminuendo linearmente fino al 50% di umidità relativa fino a 35°C. Categoria di sovratensione (IEC 60364II 4-443) Grado di contaminazione 2 Classe di protezione I Non adatto per l'uso in ambienti pericolosi. IT-30 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.1.7 Centrifuge FC5916R Modello Intervallo velocità RCF massimo Capacità massima (rotore) Intervallo di temperatura (Digitale) FC5916R 200 rpm -16000 rpm;10 rpm/impostato 26331 x g;10 x g/impostato 4X750 ml -20~40℃ Tempo di esecuzione 10 sec a 99 ore 59 min 59 sec o continuo Livello di rumore (a seconda del rotore) Densità consentita alla massima velocità Energia cinetica consentita Collegamento alla rete elettrica CA ≤ 63 ± 2 dB(A) 1,2 g/ml Fluttuazione di tensione Consumo di corrente Consumo di energia Dimensioni (L × P × A) Peso netto (senza rotore) Dimensioni di spedizione (L × P × A) Peso di spedizione (senza rotore) Condizioni ambientali (EN/IEC 61010-1) Ambiente Altitudine Temperatura ambiente Umidità relativa massima 54458 Nm 230 V ~ 120 V ~ 50/60 Hz 50/60 Hz ± 10 % ± 10 % 7,2 A 20 A 1630 W 1750 W 730 x 670 x 390 mm 28,7 x 26,4 x 15,4 pollici 118 kg 260 libbre 900 x 750 x 560 mm 40,0 x 29,5 x 22,0 pollici 137 kg 137 kg solo per uso interno Utilizzare fino ad un'altitudine di 2000 m 2°C fino a 35°C Umidità relativa max. 80% per temperature fino a 31°C, diminuendo linearmente fino al 50% di umidità relativa fino a 35°C. Categoria di sovratensione (IEC 60364II 4-443) Grado di contaminazione 2 Classe di protezione I Non adatto per l'uso in ambienti pericolosi. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 9.2 Disegni e dimensioni D H W D H W IT-31 IT-32 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Modello FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R L (mm / pollici) 362 / 14.3 408 / 16.1 408 / 16.1 446 / 17.6 723 / 28.5 540 / 21.3 730 / 28.7 P (mm / pollici) 493 / 19.4 499 / 19.7 731 / 28.8 538 / 21.2 538 / 21.2 670 / 26.4 670 / 26.4 A (mm / in.) 330 / 13.0 351 / 13.8 359 / 14.1 354 / 13.9 354 / 13.9 390 / 15.4 390 / 15.4 10. CONFORMITA Il corrispondente marchio di prodotto riporta la conformità alle normative seguenti. Marcatura Norma Questo prodotto è conforme con le normative armonizzate applicabili delle direttive Europee 2011/65/UE (RoHS), 2014/30/UE (EMC) e 2014/35/UE (LVD). La dichiarazione di Conformità Completa è disponibile online su www.ohaus.com/ce. Smaltimento In conformità alla Direttiva Europea 2012/19/UE sui Rifiuti di Apparecchiature Elettrice ed Elettroniche (RAEE) , questo dispositivo non può essere smaltito con i rifiuti domestici . Ciò vale anche per i paesi al di fuori dell'UE, come da requisiti specifici Si prega di smaltire il prodotto conformemente ai regolamenti locali al punto di raccolta specificato per le apparecchiature elettriche ed elettroniche. Se avete domande, vi invitiamo a prendere contatto con l'autorità responsabile oppure il distributore da cui avete acquistato questo dispositivo. Se questo dispositivo viene trasmesso ad altre persone (per utilizzo privato o professionale), anche il contenuto di questo regolamento deve essere trasmesso. per le istruzioni sullo smaltimento in Europa, fare riferimento a www.ohaus.com/weee. Grazie per il vostro contributo alla protezione ambientale. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11. APPENDICE TABELLA 1: PESO NETTO AMMESSO TABELLA 2: TEMPERATURE PIÙ BASSE A VELOCITÀ MAX. TABELLA 3: VELOCITÀ MAX. E VALORI RCF PER ROTORI AMMISSIBILI TABELLA 4: TEMPI DI ACCELERAZIONE E DECELERAZIONE TABELLA 5: MESSAGGI DI ERRORE TABELLA 6; CORREZIONE DEL RAGGIO TABELLA 7: MODULO DI RISANAMENTO/CERTIFICATO DI DECONTAMINAZIONE IT-33 IT-34 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.1 Table 1: PESO NETTO AMMESSO FC5916R FC5916 FC5816R FC5816 Peso netto consentito FC5718R Descrizione del rotore FC5718 Numero del rotore FC5714 Velocità massima ( x1000 rpm) 14.0 14.0 15.0 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 24x 3.4 g 15.0 16.0 16.0 16.0 30314820 Rotor, Sw ing out, 4x 250ml, ID 4.5 4x 536 g 4.5 8.0 30314821 Rotor, Angle, 6x 250ml FB, ID 8.0 6x 355 g 8.0 8.0 4.5 30314822 Rotor, Sw ing out, 4x 100ml, ID 4x 340 g 4.5 4.5 4.0 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 4x 465 g 5.0 5.0 4.5 4.5 4.5 30314824 Rotor, Sw ing out, 2x 3MTP, ID 2x 310 g 4.5 4.5 4.5 4.5 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 6x 140 g 11.0 13.5 9.0 9.0 30314826 Rotor, Angle, 6x 85ml RB, ID 11.0 6x 140 g 13.0 11.0 11.0 12.0 12.0 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 4x 140 g 12.0 12.0 15.0 15.0 30314828 Rotor, Sw ing out, 16x 50ml, ID 4x 557 g 4.5 4.5 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-35 7.5 7.5 30314829 Rotor, Angle, 10x 50ml FA, ID 9.0 10x 74 g 10.5 10.0 10.0 6.0 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 6x 72 g 6.0 6.0 12.0 12.0 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 13.0 6x 94 g 13.0 13.0 13.0 4.5 4.5 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30x 32 g 4.5 4.5 4.5 12.0 12.0 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 12.0 20x 18 g 12.0 12.0 12.0 6.0 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 12x 25 g 6.0 6.0 13.5 13.5 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 15.0 44x 3.8 g 16.0 16.0 16.0 12.0 13.0 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30x 3.4 g 14.0 15.0 15.0 14.0 14.0 15.0 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 24x 3.4 g 15.0 16.0 15.0 15.0 30314839 30314840 Rotor, Angle, 12x 1.5/2.0ml, ID Rotor, Angle, 64x 0.5ml, ID 12x 3.4 g 64x 2.3 g 18.0 18.0 13.5 13.5 15.0 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 4x 3.5 g 15.0 15.0 15.0 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID 4x 995 g 4x 708 g 4.0 4.0 8.0 8.0 IT-36 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.2 Table 2: TEMPERATURE PIÙ BASSE A VELOCITÀ MAX. Numero del rotore Descrizione del rotore 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 30314840 Rotor, Angle, 64x 0.5ml, ID 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 30553085 30553086 Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID Modello FC5718R FC5816R FC5916R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5916R FC5718R FC5816R FC5718R FC5816R FC5718R FC5816R FC5916R FC5916R FC5916R Velocità massima 15000 rpm 15000 rpm 16000 rpm 8000 rpm 4500 rpm 4500 rpm 5000 rpm 5000 rpm 4500 rpm 4500 rpm 4500 rpm 13500 rpm 13500 rpm 9000 rpm 9000 rpm 13000 rpm 12000 rpm 12000 rpm 16000 rpm 7500 rpm 7500 rpm 10500 rpm 6000 rpm 6000 rpm 12000 rpm 12000 rpm 13000 rpm 4500 rpm 4500 rpm 12000 rpm 12000 rpm 12000 rpm 6000 rpm 6000 rpm 13500 rpm 13500 rpm 16000 rpm 14000 rpm 14000 rpm 15000 rpm 15000 rpm 15000 rpm 16000 rpm 18000 rpm 18000 rpm 13500 rpm 13500 rpm 15000 rpm 15000 rpm 15000 rpm 4500 rpm 8000 rpm N-massimo 6 °C 6 °C 1 °C -1 ° C -7 ° C -7 ° C -3 ° C -3 ° C -6 ° C -6 ° C -8 ° C 17 ° C 17 ° C -5 ° C -5 ° C 8 °C 5 °C 5 °C 7 °C 9 °C 9 °C 2 °C -8 ° C -8 ° C -5 ° C -5 ° C -4 ° C -9 ° C -9 ° C 2 °C 2 °C -2 ° C -11.5 ° C -11.5 ° C 1 °C 1 °C 3 °C 9 °C 9 °C 3 °C 6 °C 6 °C 1 °C -2 ° C -2 ° C 2 °C 2 °C -1 ° C -1 ° C -2 ° C 4 °C 0 °C Tutte le indicazioni di temperatura si riferiscono ad una temperatura ambiente di 23 ° C. Superando questo valore o la radiazione solare diretta verso la centrifuga, questi valori non possono essere mantenuti. FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.3 Table 3: VELOCITÀ MAX. E VALORI RCF PER ROTORI AMMISSIBILI Numero del rotore 30304361 30314820 30314821 30314822 30314823 30314824 30314825 30314826 30314827 30314828 Descrizione del rotore Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable Rotor, Sw ing out, 4x 250ml, ID Rotor, Angle, 6x 250ml FB, ID Rotor, Sw ing out, 4x 100ml, ID Rotor, Sw ing out, 4x 100ml, ID, Sealable Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID Rotor, Angle, 4x 85ml RB, ID, Hi Rotor, Sw ing out, 16x 50ml, ID Modello Velocità massim RCF 18624 x g FC5714 RPM 14000 FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5816 4500 3780 x g FC5816R 4500 3780 x g FC5816 8000 10016 x g FC5816R 8000 10016 x g FC5916 8000 10016 x g FC5916R 8000 10016 x g FC5714 4500 3350 x g FC5718 4500 3350 x g FC5718R 4500 3350 x g FC5714 4000 2486 x g FC5718 5000 3885 x g FC5718R 5000 3885 x g FC5714 4500 2716 x g FC5718 4500 2716 x g FC5718R 4500 2716 x g FC5816 4500 2716 x g FC5816R 4500 2716 x g FC5916 4500 2716 x g FC5916R 4500 2716 x g FC5718 11000 13932 x g FC5718R 13500 20984 x g FC5718 9000 10413 x g FC5718R 9000 10413 x g FC5816 11000 15555 x g FC5816R 13000 21726 x g FC5916 11000 15555 x g FC5916R 13000 21726 x g FC5718 12000 14809 x g FC5718R 12000 14809 x g FC5816 12000 14809 x g FC5816R 12000 14809 x g FC5916 15000 23140 x g FC5916R 16000 26328 x g FC5816 4500 3735 x g FC5816R 4500 3735 x g IT-37 IT-38 30314829 30314830 30314831 30314832 30314833 30314834 30314835 30314836 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Rotor, Angle, 20x 10ml RB, ID, Hi Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable FC5718 7500 8174 x g FC5718R 7500 8174 x g FC5816 9000 11771 x g FC5816R 10500 16022 x g FC5916 10000 14532 x g FC5916R 10500 16022 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 12000 13522 x g FC5718R 12000 13522 x g FC5816 13000 15869 x g FC5816R 13000 15869 x g FC5916 13000 15869 x g FC5916R 13000 15869 x g FC5714 4500 2830 x g FC5718 4500 2830 x g FC5718R 4500 2830 x g FC5816 4500 2830 x g FC5816R 4500 2830 x g FC5718 12000 15775 x g FC5718R 12000 15775 x g FC5816 12000 15775 x g FC5816R 12000 15775 x g FC5916 12000 15775 x g FC5916R 12000 15775 x g FC5714 6000 4427 x g FC5718 6000 4427 x g FC5718R 6000 4427 x g FC5718 13500 17113 x g FC5718R 13500 17113 x g FC5816 15000 21128 x g FC5816R 16000 24039 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5714 12000 15131 x g FC5718 13000 17758 x g FC5718R 14000 20595 x g FC5916 15000 23643 x g FC5916R 15000 23643 x g FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314838 30314839 30314840 30314841 30553085 30553086 Rotor, Angle, 24x 1.5/2.0ml, ID Rotor, Angle, 12x 1.5/2.0ml, ID Rotor, Angle, 64x 0.5ml, ID Rotor, Angle, 4x 8-w PCR Strip, ID Rotor Sw ingOut 4x 750ml D99mm ID SL Rotor Angle 4x 500ml D69mm ID FC5714 14000 18624 x g FC5718 14000 18624 x g FC5718R 15000 21379 x g FC5816 15000 21379 x g FC5816R 16000 24325 x g FC5916 16000 24325 x g FC5916R 16000 24325 x g FC5718 18000 23542 x g FC5718R 18000 23542 x g FC5718 13500 16298 x g FC5718R 13500 16298 x g FC5718 15000 15343 x g FC5718R 15000 15343 x g FC5916 15000 15343 x g FC5916R 15000 15343 x g FC5916 4000 3434 x g FC5916R 4500 4346 x g FC5916 8000 10367 x g FC5916R 8000 10367 x g IT-39 IT-40 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.4 Table 4: TEMPI DI ACCELERAZIONE E DECELERAZIONE Articolo n. Descrizione del rotore Modello Accelerazione* Tempo in sec niveau 0 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable 30314820 Rotor, Sw ing out, 4x 250ml, ID 30314821 Rotor, Angle, 6x 250ml FB, ID 30314822 Rotor, Sw ing out, 4x 100ml, ID 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 30314824 Rotor, Sw ing out, 2x 3MTP, ID 30314825 Rotor, Angle, 6x 85ml RB, ID, Hi 30314826 Rotor, Angle, 6x 85ml RB, ID 30314827 Rotor, Angle, 4x 85ml RB, ID, Hi 30314828 Rotor, Sw ing out, 16x 50ml, ID 30314829 Rotor, Angle, 10x 50ml FA, ID 30314830 Rotor, Angle, 6x 50ml RB/FA, ID FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R 200 240 230 251 249 204 204 309 309 664 664 573 573 110 100 100 110 100 100 230 160 160 452 452 249 249 400 500 360 360 697 2313 463 549 300 300 506 506 448 448 311 311 380 380 753 753 480 480 101 89 90 niveau 9 25 23 25 25 20 21 21 34 34 130 130 66 66 14 15 15 14 15 15 25 20 20 43 43 27 27 40 60 40 40 85 70 48 69 60 60 60 60 50 50 34 34 63 62 115 115 60 60 11 13 12 Decelerazione* Tempo in sec niveau 0 230 240 420 610 635 421 421 458 458 2906 2906 1903 1903 170 150 150 170 150 150 340 360 360 616 616 488 488 960 1260 1050 1050 2313 1630 1654 1307 820 820 1745 1745 1251 1251 387 387 796 1069 2395 2395 1747 1747 206 239 463 niveau 9 35 16 17 26 27 30 30 36 36 92 92 84 84 19 15 15 17 15 15 26 15 15 38 38 23 23 40 50 40 40 70 76 46 67 40 40 49 49 45 45 36 36 37 36 72 72 68 68 14 11 11 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314831 Rotor, Angle, 6x 50ml RB, ID, Hi 30314832 Rotor, Angle, 30x 15ml RB/FA, ID 30314833 Rotor, Angle, 20x 10ml RB, ID, Hi 30314834 Rotor, Angle, 12x 15ml RB/FA, ID 30314835 Rotor, Angle, 44x 1.5/2.0ml, ID 30314836 Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable 30314838 Rotor, Angle, 24x 1.5/2.0ml, ID 30314839 Rotor, Angle, 12x 1.5/2.0ml, ID 30314840 Rotor, Angle, 64x 0.5ml, ID 30314841 Rotor, Angle, 4x 8-w PCR Strip, ID 30553085 Rotor Sw ingOut 4x 750ml D99mm ID SL 30553086 Rotor Angle 4x 500ml D69mm ID * Nota: accelera da 0 a Vmax; decelera da Vmax a 0. FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5816 FC5816R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5714 FC5718 FC5718R FC5916 FC5916R FC5714 FC5718 FC5718R FC5816 FC5816R FC5916 FC5916R FC5718 FC5718R FC5718 FC5718R FC5718 FC5718R FC5916 FC5916R FC5916 FC5916R FC5916 FC5916R 360 360 446 446 264 264 157 160 160 149 149 360 360 512 512 305 305 102 88 88 196 196 257 274 233 233 250 200 210 221 221 200 210 230 251 249 204 204 240 240 150 150 130 130 100 100 483 551 575 575 40 40 48 48 28 28 19 15 15 25 25 40 40 54 54 32 32 11 13 12 22 22 29 31 25 25 27 25 25 23 23 22 23 25 25 30 21 21 25 25 17 17 16 16 12 12 47 55 73 73 570 570 1323 1323 921 921 370 380 380 985 985 570 570 1439 1439 988 988 167 222 433 204 204 436 485 283 500 280 360 360 561 561 230 240 420 610 635 421 421 210 210 170 170 130 130 201 201 1287 1501 2317 2317 IT-41 30 30 49 49 32 32 20 10 10 20 20 30 30 47 47 37 37 14 11 11 18 18 28 29 30 25 34 25 30 30 30 35 16 17 26 27 30 30 20 20 12 12 12 12 12 12 49 54 82 82 IT-42 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 11.5 Tableau 5 : MESSAGGI DI ERRORE Error-No.: Descri pti on 1 I mbal ance arose 2 I mbal ance sensor i s defecti v e 4 I mbal ance swi tch has been acti vated for l onger than 5 secon ds 8 T ransponder i n the rotor i s defecti v e 11 T emperatu re sensor i s defecti v e 12 Chamber over temperature 14 Leap of speed i s too bi g between two mesaur em e nts CLOSE l i d 33 34 38 40 Open l i d whi l e motor i s runni ng Li d contact defecti v e Li d motor i s bl ocked Communi cati o n wi th frequen cy converter di strubed duri ng start 41 Communi cati o n wi th frequen cy converter di strubed duri ng stop 42 Short ci rcui t i n the frequency converter 43 Undervol tag e frequen cy converter 44 Overvol tag e frequen cy converter 45 Over temperatu re frequency converter 46 Over temperatu re motor 47 Over current frequen cy converter 48 T i meout between control uni t and frequency converter 49 Other error frequency converter 55 Overspeed 70 T i meout between control er and RS232 i nterface 99 Rotor i s not al l owed i n thi s centri fuge FALSE rotor no I nserted rotor does not exi st i n the program m Rotor i s not detected FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R IT-43 11.6 Table 6: CORREZIONE DEL RAGGIO Compatible FC5916 FC5916R Blank NO Bucket/Adapter 8.5 ● ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.3 ● ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 -1.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 16.7 ● ● 30314903 Rack, 1x 250ml D62mm FB, 2/pk 15.7 -1.0 ● ● 30314904 Rack, 1x 100ml D41mm RB, 2/pk 15.9 -0.8 ● ● 30314905 Rack, 3x 50ml D29mm RB, 2/pk 16.3 -0.4 ● ● 30314906 Rack, 3x 50ml D29mm FA, 2/pk 16.7 0.0 ● ● 30314907 Rack, 1x 50ml D34mm RB, 2/pk 15.8 -0.9 ● ● 30314908 Rack, 2x 50ml D29mm Rim, 2/pk 15.8 -0.9 ● ● 30314909 Rack, 1x 50ml D29mm Rim, 2/pk 16.1 -0.6 ● ● 30314910 Rack, 5x 25ml D24.5mm RB, 2/pk 16.1 -0.6 ● ● 30314911 Rack, 3x 25ml D24.5mm Rim, 2/pk 15.5 -1.2 ● ● 30314912 Rack, 9x 15ml D17mm RB, 2/pk 16.3 -0.4 ● ● 30314913 Rack, 7x 15ml D17mm FA, 2/pk 16.4 -0.3 ● ● 30314914 Rack, 7x 15ml D17mm Rim, 2/pk 16.1 -0.6 ● ● 30314915 Rack, 8x 10ml D16mm RB, 2/pk 16.3 -0.4 ● ● 30314916 Rack, 14x 5-7ml D13mm RB, 2/pk 15.9 -0.8 ● ● 30314917 Rack, 10x 1.6-7ml D13mm Vac, 2/pk 15.9 -0.8 ● ● 30304367 Rack, 8x 4-10ml D16mm Vac, 2/pk 16.3 -0.4 ● ● 30304368 Adapter, 9x 1.5/2.0ml D11mm, 2/pk 15.7 -1.0 ● ● Blank NO Bucket/Adapter 14.1 30304369 30314821 Rotor, Angle, 6x 250ml FB, ID ● ● ● 13.2 ● ● ● ● 30304370 Adapter, 5x 10ml D16mm RB, 2/pk 13.3 ● ● ● ● 30304371 Adapter, 4x 15ml D17mm FA, 2/pk 13.0 -1.1 ● ● ● ● 12.5 -1.6 ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● 30304372 Adapter, 2x 30ml D26mm RB, 2/pk 30304373 Adapter, 1x 50ml D29mm RB, 2/pk 12.0 -2.1 30304374 Adapter, 1x 50ml D30mm FA, 2/pk Adapter 15x 1.5/2.0ml D11mm MTBx 2 V1 11.7 -2.4 9.5 13.6 11.8 Blank Adapter 1x 250ml D60mm CBCx 3 V1 Adapter 1x 175/225ml D61mm FAx 2 V1 NO Bucket/Adapter -4.6 -0.5 -0.5 ● ● ● 30314842 Rack, 1x 100ml D46mm RB, 2/pk 14.1 -0.7 ● ● ● 30314843 Rack, 1x 85ml D38mm RB, 2/pk 14.1 -0.7 ● ● ● 30314844 Rack, 1x 50ml D34mm RB, 2/pk 14.2 -0.6 ● ● ● 30314845 Rack, 1x 50ml D29mm RB, 2/pk 14.2 -0.6 ● ● ● 30314846 Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.7 ● ● ● 30314847 Rack, 1x 50ml D29mm Rim, 2/pk 14.3 -0.5 ● ● ● 30314848 Rack, 1x 30ml D25mm Rim, 2/pk 13.8 -1.0 ● ● ● 30314849 Rack, 1x 25ml D25mm RB, 2/pk 14.4 -0.4 ● ● ● 30314850 Rack, 2x 15ml D17mm FA, 2/pk 14.8 0.0 ● ● ● 30559413 30559414 Rotor, Sw ing out, 4x 100ml, ID Adapter, 8x 1.5ml D11mm, 2/pk ● -0.9 -0.8 30559412 30314822 Correzione (cm) FC5816R Rotor, Sw ing out, 4x 250ml, ID Raggio (cm) FC5816 30314820 Descrizione della pala/adattatore FC5718R 30304361 Rotor, Angle, 24x 1.5/2.0ml, ID, Sealable Pala/Adatt atore FC5718 Descrizione del rotore FC5714 Rotore 12.8 -2.3 14.8 IT-44 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314851 Rack, 1x 15ml D17mm Rim, 2/pk 14.5 -0.3 ● ● ● 30314852 Rack, 7x 15ml D17mm RB, 2/pk 14.5 -0.3 ● ● ● 30314853 Rack, 9x 5-7ml D13mm RB, 2/pk 14.2 -0.6 ● ● ● 30314854 Rack, 10x 1.5ml D11mm, 2/pk 9.9 -4.9 ● ● ● 30314855 Rack, 9x 4.5ml D12mm FB, 2/pk 14.0 -0.8 ● ● ● 30314856 Rack, 7x 5-7ml D13mm RB, 2/pk 11.6 -3.2 ● ● ● 30314857 Rack, 7x 5-7ml D13mm Vac, 2/pk 14.2 -0.6 ● ● ● 30314858 Rack, 5x 4-10ml D16mm Sar, 2/pk 14.3 -0.5 ● ● ● Blank 14.6 ● ● ● 30314861 NO Bucket/Adapter Bucket, 100ml w o Cap, Sealable, 2/pk Bucket, 100ml w / Cap, Sealable, 2/pk 30314864 Rack, 1x 100ml D40mm RB, 2/pk 30314865 30314860 30314823 Rotor, Sw ing out, 4x 100ml, ID, Sealable 14.2 -0.4 ● ● ● 14.2 -0.4 ● ● ● 13.8 -0.8 ● ● ● Rack, 1x 85ml D38mm RB, 2/pk 14.0 -0.6 ● ● ● 30314866 Rack, 1x 50ml D34mm RB, 2/pk 14.0 -0.6 ● ● ● 30314867 Rack, 1x 50ml D29mm RB, 2/pk 14.0 -0.6 ● ● ● 30314868 Rack, 1x 50ml D29mm FA, 2/pk 14.1 -0.5 ● ● ● 30314869 Rack, 1x 50ml D29mm Rim, 2/pk 13.9 -0.7 ● ● ● 30314870 Rack, 1x 30ml D25mm Rim, 2/pk 13.1 -1.5 ● ● ● 30314871 Rack, 1x 25ml D25mm RB, 2/pk 14.0 -0.6 ● ● ● 30314872 Rack, 1x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314873 Rack, 2x 15ml D17mm FA, 2/pk 14.1 -0.5 ● ● ● 30314874 Rack, 2x 15ml D17mm Rim, 2/pk 14.0 -0.6 ● ● ● 30314875 Rack, 4x 15ml D17mm RB, 2/pk 14.0 -0.6 ● ● ● 30314876 Rack, 7x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314877 Rack, 5x 1.5ml D11mm, 2/pk 14.0 -0.6 ● ● ● 30314878 Rack, 3x 10ml D17mm Sar, 2/pk 14.0 -0.6 ● ● ● 30314879 Rack, 4x 5-7ml D13mm RB, 2/pk 14.0 -0.6 ● ● ● 30314880 14.0 -0.6 ● ● ● 14.6 0.0 ● ● ● 14.6 0.0 ● ● ● 30314883 Rack, 4x 10ml D17mm Vac, 2/pk Bucket, 2x 50ml D29mm FA, w / Rack, 2/pk Bucket, 3x 15ml D17mm FA, w / Rack, 2/pk Cy to, Rectangular Bucket, 2/pk 6.3 -8.3 ● ● ● Blank NO Bucket/Adapter 12.0 ● ● ● ● ● ● ● 30314891 Adapter, 48x 1.5/2.0ml D11mm, 2/pk 12.0 ● ● ● ● ● ● ● Blank NO Bucket/Adapter 10.3 ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 9.6 -0.7 ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 9.6 -0.7 ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 10.0 -0.3 ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 9.8 -0.5 ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 9.3 -1.0 ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 10.3 0.0 ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 9.5 -0.8 ● ● Blank NO Bucket/Adapter 11.5 ● ● ● ● ● ● 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 10.9 -0.6 ● ● ● ● ● ● 30314896 Adapter, 1x 50ml D29mm FA, 2/pk 10.6 -0.9 ● ● ● ● ● ● 30314897 Adapter, 1x 15ml D17mm FA, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 11.1 -0.4 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 10.4 -1.1 ● ● ● ● ● ● 30314881 30314882 30314824 30314825 30314826 Rotor, Sw ing out, 2x 3MTP, ID Rotor, Angle, 6x 85ml RB, ID, Hi Rotor, Angle, 6x 85ml RB, ID 0.0 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R 30314827 30314828 30314829 30314830 30314831 30314832 30314834 30314835 30314836 30314838 30314839 Rotor, Angle, 4x 85ml RB, ID, Hi Rotor, Sw ing out, 16x 50ml, ID Rotor, Angle, 10x 50ml FA, ID Rotor, Angle, 6x 50ml RB/FA, ID Rotor, Angle, 6x 50ml RB, ID, Hi Rotor, Angle, 30x 15ml RB/FA, ID Rotor, Angle, 12x 15ml RB/FA, ID Rotor, Angle, 44x 1.5/2.0ml, ID Rotor, Angle, 30x 1.5/2.0ml, ID, Sealable Rotor, Angle, 24x 1.5/2.0ml, ID Rotor, Angle, 12x 1.5/2.0ml, ID Blank NO Bucket/Adapter 9.2 30314893 Adapter, 1x 15ml D17mm RB, 2/pk 8.5 30314894 Adapter, 1x 30ml D25mm RB, 2/pk 8.4 30314895 Adapter, 1x 50ml D29mm RB, 2/pk 30314896 30314897 IT-45 ● ● ● ● ● ● -0.7 ● ● ● ● ● ● -0.8 ● ● ● ● ● ● 8.9 -0.3 ● ● ● ● ● ● Adapter, 1x 50ml D29mm FA, 2/pk 8.6 -0.6 ● ● ● ● ● ● Adapter, 1x 15ml D17mm FA, 2/pk 8.3 -0.9 ● ● ● ● ● ● 30314898 Adapter, 2x 15ml D17mm RB, 2/pk 7.5 -1.7 ● ● ● ● ● ● 30314899 Adapter, 1x 16ml D18mm RB, 2/pk 8.3 -0.9 ● ● ● ● ● ● Blank NO Bucket/Adapter 16.5 ● ● 30304375 Rack, 4x 50ml D29mm FA, 2/pk 16.5 0.0 ● ● 30314583 Rack, 10x 15ml D17mm FA, 2/pk 16.5 0.0 Blank NO Bucket/Adapter 13.0 30130894 Adapter, 1x 15ml D17mm RB, 2/pk 12.8 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk Blank NO Bucket/Adapter 11.0 30130891 Adapter, 1x 30ml D26mm, 2/pk 10.7 -0.3 30130892 Adapter, 1x 16ml D18mm, 2/pk 10.3 30130893 Adapter, 1x 15ml D17mm, 2/pk 10.6 30130894 Adapter, 1x 15ml D17mm RB, 2/pk 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 30130886 ● ● ● ● ● ● ● ● -0.2 ● ● ● ● ● ● 12.2 -0.8 ● ● ● ● ● ● 10.4 -2.6 ● ● ● ● ● ● 8.9 -4.1 ● ● ● ● ● ● ● ● ● ● ● ● -0.7 ● ● ● -0.4 ● ● ● 10.6 -0.4 ● ● ● 10.2 -0.8 ● ● ● 8.3 -2.7 ● ● ● Adapter, 1x 1.5/2.0ml D11mm, 6/pk 6.7 -4.3 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130891 Adapter, 1x 30ml D26mm, 2/pk 8.2 -0.2 ● ● ● ● ● ● 30130892 Adapter, 1x 16ml D18mm, 2/pk 7.9 -0.5 ● ● ● ● ● ● 30130893 Adapter, 1x 15ml D17mm, 2/pk 8.0 -0.4 ● ● ● ● ● ● 30314892 Adapter, 1x 15ml D17.5mm FA, 2/pk 7.7 -0.7 ● ● ● ● ● ● Blank NO Bucket/Adapter 12.5 ● ● ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 12.2 -0.3 ● ● ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 10.5 -2.0 ● ● ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 9.0 -3.5 ● ● ● ● ● Blank NO Bucket/Adapter 11.0 ● ● ● 30130889 Adapter, 1x 7ml D13.5mm RB, 2/pk 10.6 -0.4 ● ● ● 30130890 Adapter, 1x 5ml D13.5mm, 2/pk 9.1 -1.9 ● ● ● 30130886 Adapter, 1x 1.5/2.0ml D11mm, 6/pk 7.7 -3.4 ● ● ● Blank NO Bucket/Adapter 8.4 ● ● ● ● ● ● 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 -0.2 ● ● ● ● ● ● 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.7 -0.7 ● ● Blank NO Bucket/Adapter 9.4 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 9.1 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 8.4 Blank NO Bucket/Adapter 8.5 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 8.2 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 7.5 Blank NO Bucket/Adapter 6.5 30130884 Adapter, 1x 0.2/0.4ml D6mm, 6/pk 6.3 30130885 Adapter, 1x 0.5ml D8mm, 6/pk 5.6 ● ● ● ● ● ● ● ● ● -0.3 ● ● ● ● ● -1.0 ● ● ● ● ● ● ● ● ● ● ● ● -0.3 ● ● ● ● ● ● ● -1.0 ● ● ● ● ● ● ● ● ● -0.2 ● ● -0.9 ● ● IT-46 FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Blank Adapter, 1x 1.5ml for ov er16000x g, 6/pk NO Bucket/Adapter 19.3 ● ● 30553138 Adapter 32x 7ml D13mm RBx 2 V1 18.7 -0.6 ● ● 30553135 Adapter 18x 15ml D17mm RBx 2 V1 18.8 -0.5 ● ● 30553132 Adapter 8x 30ml D26mm RBx 2 V1 19.1 -0.2 ● ● 30553131 Adapter 7x 50ml D29mm RBx 2 V1 19.1 -0.2 ● ● 30553127 Adapter 3x 100ml D41mm RBx 2 V1 19.1 -0.2 ● ● 30553125 18.9 -0.4 ● ● 19.1 -0.2 ● ● 30553128 Adapter 1x 250ml D62mm FBx 2 V1 Adapter 1x 225/190ml D61mm FAx 2 V1 Adapter 5x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553130 Adapter 7x 50ml D29mm FAx 2 V1 19.1 -0.2 ● ● 30553133 Adapter 14x 15ml D17mm FAx 2 V1 19.2 -0.1 ● ● 30553134 Adapter 17x 15ml D17mm FAx 2 V1 Adapter 18x 5/10ml D16mm Vac/Sarx 2 V1 Adapter 38x 1.1/1.4ml D8mm Sarx 2 V1 19.0 -0.3 ● ● 19.0 -0.3 ● ● 18.9 -0.4 ● ● Adapter 21x 10ml D16mm RBx 2 V1 Adapter 27x 1.8/7ml D13mm Vac/Sarx 2 V1 NO Bucket/Adapter 18.9 -0.4 ● ● 18.8 -0.5 ● ● ● ● 13.7 -0.8 ● ● 13.4 -1.1 ● ● 12.4 -2.1 ● ● 12.6 -1.9 ● ● 30559421 Adapter 1x 250ml D61.5mm FBx 2 V1 Adapter 1x 175/225ml D61.5mm FAx 2 V1 Adapter 1x 50ml D29.5mm FAx 2 V1 Adapter 1x 250ml D60.5mm CBCx 2 V1 Adapter 4x 30ml D25.5mm RBx 2 V1 13.8 -0.7 ● ● 30559420 Adapter 3x 50ml D29mm RBx 2 V1 14.3 -0.2 ● ● 30559422 Adapter 6x 15ml D17mm FAx 2 V1 13.8 -0.7 ● ● 30314900 30553085 Rotor Sw ingOut 4x 750ml D99mm ID SL 30553126 30553140 30559377 30553136 30553139 Blank 30564850 30559417 30553086 Rotor Angle 4x 500ml D69mm ID 30559419 30559416 6.4 -0.1 14.5 ● ● IT-47 11.7 Tabella 7: Modulo di risanamento/Certificato di decontaminazione Allegare il presente modulo a tutti i ritorni di apparecchiature e gruppi La dichiarazione completa della decontaminazione è un requisito per la ricezione e l'ulteriore elaborazione del ritorno. Se non si allega la spiegazione corrispondente, eseguiremo la decontaminazione a vostre spese. Si prega di compilare in stampatello  FrontierTM Centrifuga FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R Cognome ___________________________________________ Organizzazione/Società ________________________________________ Via: ______________________________________________________ C.A.P. _____________ luogo ___________________ Tel.: _____________ Fax:____________________ E-mail: _____________________________________ Pos. Crowd Decontaminated object Serial number Description / Comment 1 2 3 4 Gli elementi elencati qui sopra sono in contatto con le sostanze seguenti? Soluzioni acquose dannose per la salute, soluzioni tampone, acidi, alcali....  Sì  No Agenti potenzialmente infettivi …………………………………………………………… Sì  No Reagenti organici e solvente ……………………………………………………….. ..  Sì  No Sostanze radioattive: ……………………………………………. α.. β.. γ…  Sì  No Proteine dannose per la salute ………………………………………………………….. Sì  No DNA: ………………………………………………………………………………………  Sì  No Queste sostanze hanno raggiunto l'apparecchiatura/il gruppo ? …………………  Sì  No Se sì, quali : ________________________________________________________________ Descrizione delle misure per la decontaminazione delle parti elencate: Confermo l'appropriata decontaminazione Società/Servizio.____________ Luogo e Data: ____________________________________ Firma della persona autorizzata:________________________________________ Ohaus Corporation 7 Campus Drive Suite 310 Parsippany, NJ 07054 USA With offices worldwide / Con oficinas en todo el mundo/ Avec des bureaux dans le monde entier/ Mit Niederlassungen weltweit/ Con uffici in tutto il mondo http://www.ohaus.com *30323917* P/N 30323917 C © 2020 OHAUS Corporation, all rights reserved / todos los derechos reservados / tous droits réservés / Alle Rechte vorbehalten / Tutti i diritti riservati Printed in China / Impreso en China/ Imprimé en Chine / Gedruckt in China / Stampato in Cina
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272

Ohaus FC5916 120V Manual de usuario

Tipo
Manual de usuario