Ohaus ST5000-F Manual de usuario

Tipo
Manual de usuario

Este manual también es adecuado para

STARTER 5000
Bench pH Meter
Instruction Manual
STARTER 5000
Medidor de pH de banco
Manual de instrucciones
STARTER 5000
PH-mètre de laboratoire
Manuel d’instructions
TABLE OF CONTENTS
1 INTRODUCTION ................................................................ 1
1.1 Definition of Signal Warnings and Symbols ................................. 1
1.2 Safety Precautions ........................................................................... 2
2 INSTALLATION ................................................................. 3
2.1 Package contents ............................................................................. 3
2.2 Installing the stand-alone electrode holder ................................... 4
2.3 Installing the power adapter ........................................................... 4
2.4 Connect the pH electrodes ............................................................. 5
3 DISPLAY AND ICONS ....................................................... 5
3.1 Turn on the meter ............................................................................. 5
3.2 Turn off the meter............................................................................. 6
3.3 Icons .................................................................................................. 6
3.4 Menu directory ................................................................................. 9
4 CALIBRATION, MEASUREMENT AND PRINT .............. 10
4.1 Calibration ...................................................................................... 10
4.1.1 Buffer group ..................................................................................... 10
4.1.2 Performing calibration ....................................................................... 11
4.2 Sample measurement .................................................................... 12
4.2.1 pH measurement ............................................................................. 12
4.2.2 mV and Rel.mV measurement ......................................................... 12
4.3 Temperature measurement ........................................................... 12
4.4 Printing ............................................................................................ 13
5 SETUP ............................................................................. 14
5.1 System Setting ............................................................................... 14
5.1.1 System setting ►Save Mode .......................................................... 14
5.1.2 System Setting ►Output ................................................................. 14
5.1.3 System Setting ► Restore Factory Setup ....................................... 14
5.1.4 System Setting ► Service Password .............................................. 14
5.2 Measurement Setting ..................................................................... 15
5.2.1 Measurement Setting ►Endpoint Mode .......................................... 15
5.2.2 Measurement Setting ►GLP Measurement Mode .......................... 15
5.2.3 Measurement Setting ►Continuous Measure ................................. 15
5.3 Calibration Setting ......................................................................... 16
5.3.1 Calibration Setting ►Buffer Group .................................................. 16
5.3.2 Calibration Setting Calibration Reminder ..................................... 16
5.4 Data Log .......................................................................................... 16
6 MAINTENANCE ............................................................... 17
6.1 Error message ................................................................................ 17
6.2 Meter maintenance ........................................................................ 17
6.3 Electrode maintenance .................................................................. 17
6.4 Information ..................................................................................... 17
7 TECHNICAL DATA .......................................................... 18
7.1 Specifications ................................................................................. 18
8 COMPLIANCE ................................................................. 19
9 BUFFER GROUPS .......................................................... 20
STARTER 5000 Bench pH Meter EN-1
1 INTRODUCTION
Thank you for choosing this OHAUS product.
Please read the manual completely before using the STARTER 5000 bench pH meter to avoid
incorrect operation.
STARTER 5000 is a 0.001pH bench meter with a color touch-screen LCD. It’s a new generation of
intuitive and easy-to-use bench meters for reliable measurements.
A few of the many existing features are:
Bi-lingual graphical user interface
Auto, manual time endpoint which can assist users to freeze the stable reading value to
meet different customer requirements
8 buffer groups (US, EU, NIST and JJG) with 1 self-defined buffer
IP54 water and dust protection
1.1 Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and
warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument,
malfunctions and false results.
Signal Words
WARNING For a hazardous situation with medium risk, possibly resulting in
injuries or death if not avoided.
CAUTION For a hazardous situation with low risk, resulting in damage to the
device or the property or in loss of data, or injuries if not avoided.
Attention For important information about the product
Note For useful information about the product
Warning Symbols
General Hazard Electrical Shock Hazard
Explosion Hazard Corrosive Hazard
Alternating Current Direct Current
EN-2 STARTER 5000 Bench pH Meter
1.2 Safety Precautions
CAUTION Read all safety warnings before installing, making connections, or
servicing this equipment. Failure to comply with these warnings could result in
personal injury and/or property damage. Retain all instructions for future reference.
Verify that the AC adapter's input voltage range and the plug type are compatible with the
local AC power supply.
Make sure that the power cord does not pose a potential obstacle or tripping hazard.
Use the equipment only in dry locations.
Dry off any liquid spills immediately. The instrument is not watertight.
Use only approved accessories and peripherals.
Operate the equipment only under ambient conditions specified in these instructions.
Disconnect the equipment from the power supply when cleaning.
Do not operate the equipment in hazardous or unstable environments.
Service should only be performed by authorized personnel.
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
WARNING Do not use this instrument in an environment subject to explosion
hazards. The housing of the instrument is not gas tight (explosion hazard due to
spark formation, corrosion caused by the ingress of gases).
CAUTION When using chemicals and solvents, comply with the instructions of the
producer and the general lab safety rules.
STARTER 5000 Bench pH Meter EN-3
2 INSTALLATION
Carefully unpack the box.
2.1 Package contents
The model ST5000-B (basic package) has the following items:
In addition to ST5000-B content, the ST5000-F package also includes the following:
Each pH buffer powder should be dissolved in 250ml pure water or deionized water in a
volumetric flask. You also can order the bottled buffer solution from Ohaus.
More electrodes, solutions, accessories and spare parts, please contact Ohaus for more details.
Model
Description
P/N
ST350
3-in-1 glass refillable pH Electrode
30129354
ST310
3-in-1 plastic refillable pH Electrode
83033965
ST210
2-in-1 plastic refillable pH Electrode
83033966
ST320
3-in-1 plastic gel pH Electrode(no need to fill)
83033967
ST260
2-in 1 glass double salt bridge pH Electrode
30129357
ST230
2-in-1 glass muddy sample pH Electrode
83033968
STSURF
2-in-1 flat surface pH Electrode
30129470
STMICRO5
2-in-1 glass micro sample pH Electrode
30087566
STMICRO8
2-in-1 glass micro sample pH Electrode
30087569
STTEMP30
Temperature Electrode
83033970
STORP1
Gel plastic ORP electrode
30038555
STORP2
Refillable glass ORP electrode
30038553
ST5000-B
Units
ST5000 meter
1
ST5000 In use cover
1
Stand-alone electrode holder
1 set
4GB USB flash drive
1
Buffer powder set (4.01; 7.00; 10.00)
1 set
Power supply
1 set
ST350 3-in-1 glass refillable pH electrode
1
Buffer powder set (4.01; 7.00; 10.0)
83033971
Buffer pH1.68 250ml
30100424
Buffer pH4.01 250ml
30100425
Buffer pH7.00 250ml
30100427
Buffer pH10.01 250ml
30100429
Buffer pH12.45 250ml
30100440
pH electrode Reference Electrolyte
30059255
pH electrode protection solution
30059256
Electrode holder stand alone
30058733
ST5000 In use cover
30129897
SP Cable Adapter RS232 F-M SF40A (for ST5000)
30059316
EN-4 STARTER 5000 Bench pH Meter
2.2 Installing the stand-alone electrode holder
Install the electrode arm on the base:
After adjusting the tension knob to some extent, the upper arm can be move up and down.
When the pH electrode is placed on the arm and calibration or measurement is being
performed, the storage bottle of pH electrode can be placed into the base hole for easy
storage:
2.3 Installing the power adapter
Insert the adapter plug into the power adapter slot.
CAUTION Only use an adapter plug that is compatible with the mains power
receptacle. Make sure that the adapter plug is fully seated in the power adapter slot
before plugging it into the mains power receptacle.
STARTER 5000 Bench pH Meter EN-5
2.4 Connect the pH electrodes
There are 2 sockets for the electrode. “pH” socket (BNC); “TEMP.” Socket (Cinch).
For ST350 3-in-1 electrode, connect both the BNC and Cinch socket. For 2-in-1 pH electrodes
(e.g. ST230), only connect the BNC.
3 DISPLAY AND ICONS
3.1 Turn on the meter
STARTER 5000 is a color touch screen 0.001pH bench meter. It does not have any buttons.
After connecting with power, the meter will turn on, if meter in screen protection, you can tap the
screen to turn on the screen.
Note: If customer did not operate the meter for more than 6 hours, the meter will completely shut
off automatically. You cannot turn on the meter by tapping the screen any more. You need to
reconnect power to turn on the meter.
Select the language and then press to enter the main interface.
Main Interface
EN-6 STARTER 5000 Bench pH Meter
3.2 Turn off the meter
To turn off the meter, please tap the turn-off button in the top-left of main interface to enter
the turn-off interface. Then tap to turn-off the meter or tap to go back to
main interface again.
Turn-off interface
3.3 Icons
pH measurement mode. Shift between mV and Rel.mV mode or
Endpoint icons, auto-endpoint, manual-endpoint and time-endpoint.
When in manual-endpoint, if the signal is stable, the meter will show the stable signal
icon. Tap the stop icon to reach manual endpoint.
STARTER 5000 Bench pH Meter EN-7
Information, measure and calibrate tap area
The information area is also the shortcut to enter the setting interface:
"Endpoint:"
Tap this area to enter the interface to set the endpoint mode. Available
modes are Auto, Manual and Time endpoint. The current endpoint mode is
displayed.
"Sensor ID:"
Tap this area to enter the interface to select, set or change the sensor ID.
The current sensor ID is displayed.
"Sensor Cond:"
Tap this area to enter the sensor calibration data interface. The current
sensor performance icon (judged by calibration data) is displayed.
There are 3 sensor condition icons:
Slope: more than 95%
and offset: ± (0-15) mV
Electrode condition is good
Slope: 90-95%
or offset: ± (15-35) mV
Electrode condition is
acceptable
Slope: less than 90%
or offset: ± (35-60) mV
Electrode condition is not
good or needs cleaning
Calibration reminder icon. Set the time you want the meter to alert you to do
calibration from last calibration. When the time is reached, this icon will be displayed
and blink to remind that it is time to perform a calibration.
Buffers:
This area displays the current buffer group, tap this area to enter the interface to
select the buffer group.
Start icon. Tap this icon and the meter will begin measurement, the signal
icon will blink. The start icon area will show the stop icon
Calibration icon. Tap this icon to start calibration.
Save icon. After saving the current measurement result (endpoint reached), this
icon color will change from black to light gray , indicating that the reading has
already been saved and cannot be saved twice.
Output icon. Output data to external device such as a printer.
If “Shut off connection” is chosen in System Setting Output, the save
icon will be displayed .Otherwise the output icon will be displayed in the
same position.
EN-8 STARTER 5000 Bench pH Meter
Return to main interface
Return to previous menu
Enter, confirm or apply current setup, finish setting
Edit
Page up
Page down
Delete
Increase (light)
Decrease (light)
Graph display (for the measurement or calibration data)
Save
Output to external device (USB flash drive, printer)
Full keyboard
Next calibration point
Current Calibration point again
End of calibration
STARTER 5000 Bench pH Meter EN-9
3.4 Menu directory
The menus of first and secondary directory for STARTER 5000:
System
Setting
Date
Time
Save Mode
Output
Screen Brightness
Baud Rate
Screen Protection
Restore Factory Setup
Instrument Information
Service Password
Measurement
Setting
Resolution
MTC Temp. Input
pH/mV/Rel.mV Mode
Continuous Measure
Endpoint Mode
ID setting
GLP Measurement Mode
Calibration
Setting
Buffer Group
Calibration Reminder
Data
Measurement Data
Calibration Data
EN-10 STARTER 5000 Bench pH Meter
4 CALIBRATION, MEASUREMENT AND PRINT
The meter display will turn off automatically according to the time set in the menu System Setting
Screen Protection LCD Auto-off. To turn on the screen again, tap the screen.(See 3.1)
A complete pH measurement process should follow the procedure:
a) Rinse the pH electrode and connect to meter
b) Prepare the buffer and calibrate the pH electrode
c) Prepare the sample
d) Conduct pH measurement of the sample
e) Record measurement data
f) Rinse the pH electrode and store
Prepare the pH electrode: the pH electrode should be rinsed with pure water before and after
usage. Check if the electrode is physically damaged. (Be careful with the glass bulb)
The pH electrode should be stored in the storage bottle. After the pH electrode has been put into
the sample or buffer solution, stir several seconds then wait 30 to 60 seconds for the signal to be
stable, and then press the button to operate (Calibration or measurement).
WARNING Do not operate the equipment in hazardous environments. The equipment
is not explosion protected.
CAUTION When using chemicals and solvents, comply with the instructions of the
chemical producer and the general lab safety rules.
4.1 Calibration
4.1.1 Buffer group
STARTER 5000 can perform 1-5 point calibrations for predefined buffer groups. For self-defined
buffer group, it can do 1-9 point calibration.
Before calibration, make sure the current buffer group is the group wanted.
When using a new pH electrode, a calibration should be performed before any measurement. If a
pH electrode is not used for a few days, a calibration should be performed before next usage.
Calibration is to display the right pH value when the meter receive the mV value signal from the
pH electrode.
Slope: the linear coefficient between mV and pH according to theoretical value (e.g. -
59.16mV/pH @ 25°C means 100% slope);
Offset: the mV value when the pH value is 7.00. (Theoretical value is 0 mV)
STARTER 5000 Bench pH Meter EN-11
STARTER 5000 will give an electrode condition icon to show the performance of the pH electrode
according to the calibration data. Normally, if we get sad face icon after a proper calibration, you
may need to replace the pH electrode with a new one.
STARTER 5000 automatically corrects for the temperature dependence of the buffer pH values.
E.g. for US standard buffer group "1.68, 4.01, 7.00, 10.01" reference the following table:
For example, with calibration buffer 10.01 at 15°C , the calibration result should be 10.12 pH.
4.1.2 Performing calibration
After rinsing the pH electrode, put the electrode into the first buffer, stir for 10s then press to
start the calibration. When in calibration, the signal icon blinks. When the endpoint has been
reached, appears while disappears and current point calibration is finished, There
are now 3 options:
Quit the calibration and go back to main interface
Finish current point calibration, go to calibration result interface
Do the next point calibration
Note: During calibrating process, the ST5000 is always in Auto Endpoint Mode.
Note: The use of 3-in-1 pH electrode which include the temperature sensor or a separate
temperature electrode is recommended. When using the MTC mode, enter the correct
temperature value and keep all buffer and sample solutions at the set temperature.
5°C
1.67
4.01
7.09
10.25
10°C
1.67
4.00
7.06
10.18
15°C
1.67
4.00
7.04
10.12
20°C
1.68
4.00
7.02
10.06
25°C
1.68
4.01
7.00
10.01
30°C
1.68
4.01
6.99
9.97
35°C
1.69
4.02
6.98
9.93
40°C
1.69
4.03
6.97
9.89
45°C
1.70
4.05
6.97
9.86
50°C
1.71
4.06
6.96
9.83
EN-12 STARTER 5000 Bench pH Meter
4.2 Sample measurement
4.2.1 pH measurement
Be sure the electrode has been calibrated before measurement. After rinsing,
Place the pH electrode in the sample.
Press to start the pH measurement, when the endpoint has been reached, the current
measurement is finished.
Note: If the signal is not stable all the time, it is recommended to use manual endpoint mode to
avoid bad repeatability.
4.2.2 mV and Rel.mV measurement
Press to shift to or mode; follow the same procedure as for pH
measurement to perform mV or Rel.mV measurement.
If the meter is connected to an ORP electrode, the mV result is the value of Oxidation-Reduction
Potential value; also called Redox value.
4.3 Temperature measurement
For better accuracy, we recommend to use either a built-in or a separate temperature electrode.
If a temperature electrode is used, ATC and the sample temperature are displayed.
If the meter does not detect a temperature electrode, it automatically switches to the manual
temperature compensation mode and MTC appears. MTC temperature should be set.
Note: STARTER 5000 accepts NTC 30 kΩ temperature sensor. For ATC (Auto Temperature
Compensation) or MTC (Manual Temperature Compensation), the Temperature Compensation
only corrects for the change in the output of the electrode, not for the change in the actual solution.
That means the meter corrects the signal (mV) from the pH electrodes to get a more accurate pH
value according to the real temperature.
STARTER 5000 Bench pH Meter EN-13
4.4 Printing
STARTER 5000 can print the measurement or calibration data through RS232. After the printer
(e.g. OHAUS SF-40A) is connected to the meter properly, make the following settings(See 5.1):
please order the connect cable for ST5000 (P/N 30059316) to connect SF40A and ST5000.
System Setting Output , select Printer
System Setting Baud Rate, select 9600.
The format for the print-out measurement is:
Date Time User SampleID
6-13-2015 14:20:22 User01 Sample02
SensorID Endpoint mode MTC/ATC
Sensor02 Manual ATC
Temperature Result
25.3°C 6.999pH
The format for the print-out calibration data is:
Calibration Data
SensorID Serial Number
Sensor02 14060012
Cal. Date Cal. Time ATC/MTC
01-01-2015 14:20:33 ATC
Buffers mV Offset Slope
4.00pH 176.94mV
7.00pH 0.05mV 0.05mV 98.22%
When performing continuous measurement, the format for the print-out calibration data is:
Timing Measure Record
SensorID Serial Number
Sensor03 14060012
Date Time Temperature Measurement
01-01-2015 14:20:25 23.5°C 6.998pH
Date Time Temperature Measurement
01-01-2015 14:20:30 23.5°C 6.999pH
Date Time Temperature Measurement
01-01-2015 14:20:35 23.5°C 7.000pH
EN-14 STARTER 5000 Bench pH Meter
5 SETUP
5.1 System Setting
System setting is used to set meter parameters, the meter information could also be found in it.
Please see section 3.4 for reference.
5.1.1 System setting Save Mode
Save mode has 2 options, one is Auto Save Endpoint Reading, another is Manual Save Endpoint
Reading. Auto Save Endpoint Reading means that when the measurement reaches the endpoint
(auto, manual or time endpoint), the meter will save the endpoint reading automatically.
The Save icon will change from black to light grey ; An endpoint reading can be
saved only once.
5.1.2 System Setting Output
When selecting shut off connection, the main interface will display the save icon , or
display the Output Icon .
5.1.3 System Setting Restore Factory Setup
Restoring the factory setup includes the following:
1. Shut off GLP mode
2. Shut off Auto Save mode
3. Shut off continuous measure
4. Shut off (output) connection
5. Shut off calibration Reminder
6. The resolution is: 0.001pH, 0.01mV
7. Measurement mode: pH
8. Self buffer group will be deleted
9. Baud rate: 9600
10. Brightness of screen: 70%
11. Shut off Screen Protection
12. MTC: 25°C
13. Endpoint: Auto
14. Time endpoint: 10 Second
15. Delete all the calibration data
5.1.4 System Setting Service Password
Not used.
STARTER 5000 Bench pH Meter EN-15
5.2 Measurement Setting
5.2.1 Measurement Setting Endpoint Mode
Auto Endpoint
In this mode, the meter will reach the endpoint and lock the reading according to the following
stability criteria automatically:
0.1pH resolution: the signal of the sensor input may not change by more than 0.3mV in 6
seconds.
0.01pH resolution: the signal of the sensor input may not change by more than 0.1mV in 6
seconds.
0.001pH resolution: the signal of the sensor input may not change by more than 0.03mV in 6
seconds.
Time Endpoint
In this mode, the user needs to input the endpoint time, e.g. 15 seconds. This means after starting
the measurement for 15s, the meter will reach the endpoint.
Manual Endpoint
In this mode, the user needs to decide when the signal reading is stable, then press the stop icon
to reach manual endpoint.
5.2.2 Measurement Setting GLP Measurement Mode
GLP mode requires each measurement to have the User ID and Sample ID. If there is no sample
ID or user ID, the meter will show the interface to let user input the required data.
5.2.3 Measurement Setting Continuous Measure
During continuous measure, the user needs to input the time interval to get each reading result.
Below is the Graph of continuous measure.
The Y axis is pH value. The x axis is the Number of measurement result.
EN-16 STARTER 5000 Bench pH Meter
5.3 Calibration Setting
5.3.1 Calibration Setting Buffer Group
Before calibration, please make sure the current buffer group is correct for your buffer solutions.
User can select from the 8 predefined buffer groups.
User can also set self-defined buffer groups. The adjacent 2 calibration point difference should be
larger than 0.99pH.
5.3.2 Calibration Setting Calibration Reminder
Enter this setting to set the time, from the last calibration, after how many hours the meter should
remind to perform calibration again. The meter will show the calibration reminder icon and
blink, to remind user to perform calibration.
5.4 Data Log
Tap Data to get into the Data log interface, including measurement data and calibration data.
You can delete all the measurement data here. The only way to delete calibration data is to
restore factory setup.
The data can be output to a USB Flash Drive (4GB).
STARTER 5000 Bench pH Meter EN-17
6 MAINTENANCE
6.1 Error message
When the input value/measurement result is out of range or calibration data is not good enough,
the meter will display error information.
If the error cannot be solved by the error message displayed, please contact Ohaus sales
representative for further assistance.
6.2 Meter maintenance
The STARTER 5000 series instruments do not require any maintenance other than occasional
wiping with a damp cloth.
Attention: Follow these safety precautions when cleaning the instrument
Disconnect the instrument from the power supply
Do not open the instrument
Do not allow liquids to enter the instrument
Attention: The housing is made of acrylonitrile butadiene styrene (ABS). This material is attacked
by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage
should be immediately wiped off.
6.3 Electrode maintenance
Make sure the electrode is filled with electrolyte solution. Always store the electrode according to
the electrode instruction manual and do not allow it to dry out.
If the electrode response becomes sluggish or the slope is not good enough, try the following:
o Soak the electrode in 0.1M HCl for more than 8 hours.
o For fat or oil contaminant, degrease the membrane with cotton wool soaked in either
acetone or a soap solution.
After electrode treatment, a new calibration should be performed. If the electrode slope is still not
good, the electrode might need to be replaced.
6.4 Information
Tap the Information icon to enter the information interface. Here the user can input related
information by tapping the area with " " in the bottom left such as sensor purchase date, etc.
EN-18 STARTER 5000 Bench pH Meter
7 TECHNICAL DATA
7.1 Specifications
The technical data is valid under the following ambient conditions:
Indoor use only
Altitude: Up to 2000 m
Specified Temperature range: 5°C to 40°C
Humidity: maximum relative humidity 80 % for temperatures up to 30°C decreasing linearly to
50% relative humidity at 40°C
Mains supply voltage fluctuations: up to ±10% of the nominal voltage
Installation category II
Pollution degree: 2
Operability is assured at ambient temperatures between 5°C to 40°C
Model
STARTER 5000
Measuring range
-2.00020.000 pH
2000.002000.00 mV
-30°C …130°C
Resolution
0.1/0.01/0.001 pH
1/0.1/0.01 mV
0.1°C
Error limits
± 0.002 pH
± 0.03% FS mV
± 0.1°C
Calibration
1-9 points
8 predefined buffer group+1 self-defined buffer group
Memory
1000 measurements
10*10 calibration data
Power supply
100-240V~0.3A input, 9V 1.12A output AC Adapter
Size/weight
Approximately 210 x 170 x 80 mm / 0.65 kg
Ship dimension/weight
Approximately 372 x 310 x 185 mm / 2.30 kg
Display
Touch screen liquid crystal with backlight
Input
BNC, impedance > 3 x 10e+12 Ω
Cinch, NTC 30 kΩ
Temperature-compensation
ATC & MTC
Housing
ABS
STARTER 5000 Bench pH Meter EN-19
8 COMPLIANCE
This product conforms to the EMC Directive 2004/108/EC and the Low Voltage
Directive 2006/95/EC. The Declaration of Conformity is available online at
europe.ohaus.com/europe/en/home/support/compliance.aspx.
In conformance with the European Directive 2002/96/EC on Waste Electrical
and Electronic Equipment (WEEE) this device may not be disposed of in
domestic waste. This also applies to countries outside the EU, per their specific
requirements. The Batteries Directive 2006/66/EC introduces new requirements
from September 2008 on removability of batteries from waste equipment in EU
member States. To comply with this Directive, this device has been designed for
safe removal of the batteries at end-of-life by a waste treatment facility. Please
dispose of this product in accordance with local regulations at the collecting
point specified for electrical and electronic equipment. If you have any
questions, please contact the responsible authority or the distributor from which
you purchased this device. Should this device be passed on to other parties (for
private or professional use), the content of this regulation must also be related.
Disposal instructions in Europe are available online at
europe.ohaus.com/europe/en/home/support/weee.aspx. Thank you for your
contribution to environmental protection.
FCC Note: This equipment has been tested and found to comply with the limits for a Class
A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference in which case the user will be required to correct
the interference at his own expense.
Industry Canada Note
This Class A digital apparatus complies with Canadian ICES-001
ISO 9001 Registration
In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO 9001 by
Bureau Veritus Quality International (BVQI), confirming that the OHAUS quality
management system is compliant with the ISO 9001 standards requirements. On June 21,
2012, OHAUS Corporation, USA, was re-registered to the ISO 9001:2008 standard.
EN-20 STARTER 5000 Bench pH Meter
9 BUFFER GROUPS
STARTER 5000 automatically correct for the temperature dependence of the buffer group pH value
given in the following tables. Please choose the buffer group you want before calibration.
JJG119 ((Ref. 25°C , 0.01pH)
Temp °C 1.68 4.00 6.86 9.18 12.46
5 1.67 4.00 6.95 9.39 13.21
10 1.67 4.00 6.92 9.33 13.01
15 1.67 4.00 6.90 9.28 12.82
20 1.68 4.00 6.88 9.23 12.64
25 1.68 4.00 6.86 9.18 12.46
30 1.68 4.01 6.85 9.14 12.29
35 1.69 4.02 6.84 9.11 12.13
40 1.69 4.03 6.84 9.07 11.98
45 1.70 4.04 6.83 9.04 11.83
50 1.71 4.06 6.83 9.02 11.70
US standard (Ref. 25°C )
Temp °C 1.68 4.01 7.00 10.01
5 1.67 4.01 7.09 10.25
10 1.67 4.00 7.06 10.18
15 1.67 4.00 7.04 10.12
20 1.68 4.00 7.02 10.06
25 1.68 4.01 7.00 10.01
30 1.68 4.01 6.99 9.97
35 1.69 4.02 6.98 9.93
40 1.69 4.03 6.97 9.89
45 1.70 4.05 6.97 9.86
50 1.71 4.06 6.96 9.83
Europe standard (Ref. 25°C )
Temp°C 2.00 4.01 7.00 9.21 11.00
5 2.02 4.01 7.09 9.45 11.72
10 2.01 4.00 7.06 9.38 11.54
15 2.00 4.00 7.04 9.32 11.36
20 2.00 4.00 7.02 9.26 11.18
25 2.00 4.01 7.00 9.21 11.00
30 1.99 4.01 6.99 9.16 10.82
35 1.99 4.02 6.98 9.11 10.64
40 1.98 4.03 6.97 9.06 10.46
45 1.98 4.04 6.97 9.03 10.28
50 1.98 4.06 6.97 8.99 10.10
STARTER 5000 Bench pH Meter EN-21
MERCK (Ref. 20°C )
Temp °C 2.00 4.00 7.00 9.00 12.00
5 2.01 4.04 7.07 9.16 12.41
10 2.01 4.02 7.05 9.11 12.26
15 2.00 4.01 7.02 9.05 12.10
20 2.00 4.00 7.00 9.00 12.00
25 2.00 4.01 6.98 8.95 11.88
30 2.00 4.01 6.98 8.91 11.72
35 2.00 4.01 6.96 8.88 11.67
40 2.00 4.01 6.95 8.85 11.54
45 2.00 4.01 6.95 8.82 11.44
50 2.00 4.00 6.95 8.79 11.33
DIN (19267) (Ref. 25°C )
Temp °C 1.09 4.65 6.79 9.23 12.75
5 1.08 4.67 6.87 9.43 13.63
10 1.09 4.66 6.84 9.37 13.37
15 1.09 4.66 6.82 9.32 13.16
20 1.09 4.65 6.80 9.27 12.96
25 1.09 4.65 6.79 9.23 12.75
30 1.10 4.65 6.78 9.18 12.61
35 1.10 4.65 6.77 9.13 12.45
40 1.10 4.66 6.76 9.09 12.29
45 1.10 4.67 6.76 9.04 12.09
50 1.11 4.68 6.76 9.00 11.98
JIS Z 8802 (Ref. 25°C )
Temp °C 1.679 4.008 6.865 9.180
5 1.668 3.999 6.951 9.395
10 1.670 3.998 6.923 9.332
15 1.672 3.999 6.900 9.276
20 1.675 4.002 6.881 9.225
25 1.679 4.008 6.865 9.180
30 1.683 4.015 6.853 9.139
35 1.688 4.024 6.844 9.102
40 1.694 4.035 6.838 9.068
45 1.700 4.047 6.834 9.038
50 1.707 4.060 6.833 9.011
DIN(19266) / NIST (Ref. 25°C )
Temp °C 1.680 4.008 6.865 9.183 12.454
5 1.668 4.004 6.950 9.392 13.207
10 1.670 4.001 6.922 9.331 13.003
15 1.672 4.001 6.900 9.277 12.810
20 1.676 4.003 6.880 9.228 12.627
25 1.680 4.008 6.865 9.183 12.454
30 1.685 4.015 6.853 9.144 12.289
35 1.691 4.026 6.845 9.110 12.133
40 1.697 4.036 6.837 9.076 11.984
45 1.704 4.049 6.834 9.046 11.841
50 1.712 4.064 6.833 9.018 11.705
EN-22 STARTER 5000 Bench pH Meter
JJG119 (Ref. 25°C , 0.001pH)
Temp °C 1.680 4.003 6.864 9.182 12460
5 1.669 3.999 6.949 9.391 13.210
10 1.671 3.996 6.921 9.330 13.011
15 1.673 3.996 6.898 9.276 12.820
20 1.676 3.998 6.879 9.226 12.637
25 1.680 4.003 6.864 9.182 12.460
30 1.684 4.010 6.852 9.142 12.292
35 1.688 4.019 6.844 9.105 12.130
40 1.694 4.029 6.838 9.072 11.975
45 1.700 4.042 6.834 9.042 11.828
50 1.706 4.055 6.833 9.015 11.697
STARTER 5000 Bench pH Meter EN-23
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of
delivery through the duration of the warranty period. During the warranty period Ohaus will repair,
or, at its option, replace any component(s) that proves to be defective at no charge, provided that
the product is returned, freight prepaid, to Ohaus.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to
radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or
as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty
registration card, the warranty period shall begin on the date of shipment to the authorized dealer.
No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not
be liable for any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or
your local Ohaus dealer for further details.
TABLA DE CONTENIDO
1 INTRODUCCIÓN ............................................................... 1
1.1 Definición de advertencias y símbolos de señales ...................... 1
1.2 Precauciones de seguridad ............................................................ 2
2 INSTALACIÓN ................................................................... 3
2.1 Contenido del paquete .................................................................... 3
2.2 Instalación del portaelectrodos independiente ............................ 4
2.3 Instalación del adaptador de corriente .......................................... 4
2.4 Conexión de los electrodos de pH ................................................. 5
3 PANTALLA E ÍCONOS ...................................................... 5
3.1 Encendido del medidor ................................................................... 5
3.2 Apagado del medidor ...................................................................... 6
3.3 Íconos ................................................................................................ 6
3.4 Directorio del menú ......................................................................... 9
4 CALIBRACIÓN, MEDICIÓN E IMPRESIÓN .................... 10
4.1 Calibración...................................................................................... 10
4.1.1 Grupo de amortiguador .................................................................... 10
4.1.2 Calibración ........................................................................................ 11
4.2 Medición de muestras ................................................................... 12
4.2.1 Medición del pH ............................................................................... 12
4.2.2 Medición de mV y Rel.mV ............................................................... 12
4.3 Medición de temperatura .............................................................. 12
4.4 Impresión ........................................................................................ 13
5 CONFIGURACIÓN .......................................................... 14
5.1 Configuración del sistema ............................................................ 14
5.1.1 System setting (Configuración del sistema) Save Mode
(Modo Guardar) ................................................................................ 14
5.1.2 System Setting (Configuración del sistema) ►Output (Salida) ......... 4
5.1.3 System Setting (Configuración del sistema) Restore Factory
Setup (Restaurar configuración de fábrica) ..................................... 14
5.1.4 System Setting (Configuración del sistema) ► Service Password
(Contraseña de servicio) .................................................................. 14
5.2 Configuración de medición ........................................................... 15
5.2.1 Measurement Setting (Configuración de medición) Endpoint Mode
(Modo de punto final) ....................................................................... 15
5.2.2 Measurement Setting (Configuración de medición) ►GLP
Measurement Mode (Modo de medición GLP) ................................ 15
5.2.3 Measurement Setting (Configuración de medición) ►Continuous
Measure (Medición continua) ........................................................... 15
5.3 Configuración de calibración ....................................................... 16
5.3.1 Calibration Setting (Configuración de calibración) ►Buffer Group
(Grupo de amortiguador)................................................................. 16
5.3.2 Calibration Setting (Configuración de calibración) ►Calibration
Reminder (Recordatorio de calibracn) .......................................... 16
5.4 Registro de datos ........................................................................... 16
6 MANTENIMIENTO ........................................................... 17
6.1 Mensaje de error ............................................................................ 17
6.2 Mantenimiento del medidor .......................................................... 17
6.3 Mantenimiento del electrodo ........................................................ 17
6.4 Información..................................................................................... 17
7 DATOS TÉCNICOS ......................................................... 18
7.1 Especificaciones ............................................................................ 18
8 CONFORMIDAD NORMATIVA ........................................ 19
9 GRUPOS DE AMORTIGUADOR ..................................... 20
STARTER 5000 Medidor de pH de banco ES-1
1 INTRODUCCIÓN
Gracias por elegir este producto OHAUS.
Lea este manual completamente antes de usar el medidor de pH de banco STARTER 5000 para
evitar su operación incorrecta.
El STARTER 5000 es un medidor de pH de 0.001 con una pantalla LCD ctil a color. Es una nueva
generación de medidores de banco intuitivos y ciles de usar para realizar mediciones
confiables.
Algunas de las tantas características actuales son:
Interfaz de usuario gráfica bilingüe
Punto final de tiempo automático o manual que puede ayudar a los usuarios a congelar el
valor de lectura estable para atender diferentes requerimientos del cliente
8 grupos de amortiguador (US, EU, NIST y JJG) con 1 amortiguador autodefinido
Protección contra el ingreso de agua y polvo IP54
1.1 Definición de advertencias y símbolos de señales
Las notas de seguridad están marcadas con palabras de sales y mbolos de advertencia.
Estas muestran temas de seguridad y advertencias. Ignorar las notas de seguridad puede
conducir a lesiones físicas, daños a los instrumentos, mal funcionamiento y resultados falsos.
Palabras de señales
ADVERTENCIA Para una situación peligrosa con riesgo medio que posiblemente
resulte en lesiones o muerte si no se evita.
PRECAUCN Para una situación peligrosa de bajo riesgo, que resulta en daño al
dispositivo o a la propiedad, rdida de datos, o lesiones, si no se
evita.
Atención Para información importante acerca del producto
Nota Para información útil acerca del producto
mbolos de advertencia
Peligro general Riesgo de descarga eléctrica
Peligro de explosión Peligro corrosivo
Corriente alterna Corriente directa
ES-2 STARTER 5000 Medidor de pH de banco
1.2 Precauciones de seguridad
PRECAUCIÓN Lea todas las advertencias de seguridad antes de instalar, hacer
conexiones o dar servicio a este equipo. La falta de cumplimiento de estas
advertencias puede resultar en lesiones personales o daños materiales. Conserve
todas las instrucciones para referencia futura.
Verifique que el rango del voltaje de entrada del adaptador de corriente alterna y el tipo de
enchufe sean compatibles con la alimentación de corriente alterna local.
Verifique que el cordón eléctrico no represente un obstáculo potencial o riesgo de tropezarse.
Use el equipo solamente en lugares secos.
Seque de inmediato cualquier líquido que se derrame. El instrumento no es hertico al
agua.
Use solo accesorios y periféricos aprobados.
Opere el equipo solo en las condiciones ambientales especificadas en estas instrucciones.
Desconecte el equipo de la alimentación eléctrica antes de limpiarlo.
No opere el equipo en ambientes peligrosos o inestables.
El servicio debe proporcionarse solamente por personal autorizado.
Si el equipo se utiliza de una manera no especificada por el fabricante,puede reducirse la prot
ección provista por el equipo.
ADVERTENCIA No use este instrumento en un ambiente sujeto a peligros de
explosión. La caja del instrumento no es hermética al gas (peligro de explosión
debido a formación de chispas; corrosión debido al ingreso de gases).
PRECAUCN Cuando use químicos y solventes, siga las instrucciones del
fabricante y reglas de seguridad generales de laboratorio.
STARTER Medidor de pH de banco ES-3
2 INSTALACIÓN
Desempaque con cuidado la caja.
2.1 Contenido del paquete
El modelo ST5000-B (paquete básico) tiene los siguientes elementos:
Además del contenido del ST5000-B, el paquete del ST5000-F también incluye lo siguiente:
Cada polvo amortiguador de pH debe disolverse en 250 ml de agua pura o agua desionizada en
un matraz volutrico. También puede ordenar solución amortiguadora embotellada de Ohaus.
Para más electrodos, soluciones, accesorios y partes de repuesto, comuníquese a Ohaus para
s detalles.
Modelo
Descripción
mero de parte
ST350
Electrodo de pH rellenable de vidrio 3 en 1
30129354
ST310
Electrodo de pH rellenable de plástico 3 en 1
83033965
ST210
Electrodo de pH rellenable de plástico 2 en 1
83033966
ST320
Electrodo de pH de gel de plástico 3 en 1 (no se necesita
llenar)
83033967
ST260
Electrodo de pH de puente de sal doble de vidrio 2 en 1
30129357
ST230
Electrodo de pH de muestra fangosa de vidrio 2 en 1
83033968
STSURF
Electrodo de pH de superficie plana 2 en 1
30129470
STMICRO5
Electrodo de pH de muestra micro de vidrio 2 en 1
30087566
STMICRO8
Electrodo de pH de muestra micro de vidrio 2 en 1
30087569
STTEMP30
Electrodo de temperatura
83033970
STORP1
Electrodo ORP de plástico de gel
30038555
STORP2
Electrodo ORP de vidrio rellenable
30038553
ST5000-B
Unidades
Medidor ST5000
1
Cubierta en uso de ST5000
1
Portaelectrodos independiente
1 juego
Unidad USB de 4 GB
1
Juego de polvo amortiguador (4.01; 7.00; 10.00)
1 juego
Alimentación de energía
1 juego
Electrodo de pH rellenable de vidrio 3 en 1 ST350
1
Juego de polvo amortiguador (4.01; 7.00; 10.0)
83033971
Amortiguador pH 1.68 250 ml
30100424
Amortiguador pH 4.01 250 ml
30100425
Amortiguador pH 7.00 250 ml
30100427
Amortiguador pH 10.01 250 ml
30100429
Amortiguador pH 12.45 250 ml
30100440
Electrólito de referencia de electrodo de pH
30059255
Solución de protección de electrodo de pH
30059256
Portaelectrodos independiente
30058733
Cubierta en uso de ST5000
30129897
Adaptador de cable SP RS232 F-M SF40A (para ST5000)
30059316
ES-4 STARTER 5000 Medidor de pH de banco
2.2 Instalación del portaelectrodos independiente
Instale el brazo del electrodo en la base:
Después de ajustar la perilla de apriete hasta cierto punto, el brazo superior puede moverse
hacia arriba y abajo. Cuando el electrodo de pH se coloca en el brazo y se realiza la
calibración o medición, el frasco de almacenamiento del electrodo de pH puede colocarse en
el orificio de la base para fácil almacenamiento:
2.3 Instalación del adaptador de corriente
Inserte el enchufe del adaptador en la ranura del adaptador.
PRECAUCN Solamente use un enchufe de adaptador que sea compatible con
la toma de corriente de la alimentación eléctrica. Asegúrese de que el enchufe del
adaptador esté apoyado completamente sobre la ranura del adaptador antes de
conectarlo en la toma de corriente de la alimentación eléctrica.
STARTER Medidor de pH de banco ES-5
2.4 Conexión de los electrodos de pH
Hay dos conectores para el electrodo. Conector “pH” (BNC); conector “TEMP.” (Cinch).
Para el electrodo ST350 3 en 1, conecte los conectores BNC y Cinch. Para los electrodos de pH
2 en 1 (por ejemplo, el ST230), solamente conecte el BNC.
3 PANTALLA E ÍCONOS
3.1 Encendido del medidor
El STARTER 5000 es un medidor de banco de pH 0.001 de pantalla ctil a color. No tiene ningún
botón.
Después de conectar la corriente, el medidor se encenderá; si tiene protector de pantalla, toque
la pantalla para encenderla.
Nota: si el cliente no ha trabajado con el medidor durante más de 6 horas, el medidor se apagará
automáticamente. En este caso ya no podrá encender el medidor con solo tocar la pantalla. Es
necesario volver a conectar la corriente para encender el medidor.
Seleccione el idioma y presione para entrar en la interfaz principal.
Interfaz principal
ES-6 STARTER 5000 Medidor de pH de banco
3.2 Apagado del medidor
Para apagar el medidor, toque el botón de apagado en la parte superior izquierda de la
interfaz principal para entrar en apagar interfaz. Luego toque para apagar el
medidor, o toque para volver a la interfaz principal.
Apagar la interfaz
3.3 Íconos
Modo de medición de pH. Alterna entre el modo mV y Rel.mV o
Íconos de punto final, punto final automático, punto final manual y punto final de
tiempo.
Cuando está en punto final manual, si la señal es estable, el medidor mostrará el
ícono de señal estable. Toque el ícono de detener para alcanzar el punto final
manual.
STARTER 5000 Medidor de pH de banco ES-7
Área de toque de información, medición y calibración
El área de información también es el acceso rápido para entrar en la interfaz de configuración:
"Endpoint:"
Toque esta área para entrar en la interfaz para configurar el modo de punto
final. Los modos disponibles son punto final automático (Auto), Manual y de
tiempo (Time). Se muestra el modo de punto final actual.
"Sensor ID:"
Toque esta área para entrar en la interfaz para seleccionar, configurar o
cambiar la identificación del sensor. Se muestra la identificacn del sensor
actual.
"Sensor Cond:"
Toque esta área para entrar en la interfaz de datos de calibración del
sensor. Se muestra el ícono de rendimiento del sensor actual (según los
datos de calibración).
Hay 3 íconos de condición de sensor:
Pendiente: más del 95%
y desfase: ± (0-15) mV
La condición del electrodo es
buena
Pendiente: 90-95%
o desfase: ± (15-35) mV
La condición del electrodo
es aceptable
Pendiente: menos del 90%
o desfase: ± (35-60) mV
La condición del electrodo no es
buena o necesita limpieza
Ícono de recordatorio de calibración. Ajuste la hora en que desee que el medidor le
avise para hacer una calibración desde la última calibración. Cuando se llegue a la
hora, se mostrará este ícono y destellará para recordarle que es momento de
realizar una calibración.
Buffers:
Esta área muestra el grupo de amortiguador actual; toque esta área para entrar a la
interfaz y seleccionar el grupo de amortiguador.
Ícono de inicio. Toque este ícono y el medidor comenzará la medición; el ícono
de señal destellará. El área del ícono de inicio mostrará el ícono de
detener
Ícono de calibración. Toque este ícono para iniciar la calibración.
Ícono de guardar. Después de guardar el resultado de la medición actual (se
alcanzó el punto final), el color del ícono cambiará de negro a gris claro
para indicar que la lectura ya se guardó y no puede guardarse dos veces.
Ícono de salida. Envíe datos a un dispositivo externo tal como una impresora.
Si se selecciona “Shut off connection” (Apagar conexión) en System Setting
(Configuración del sistema) ►Output (Salida), se mostrará el ícono de
guardar . De lo contrario, el ícono de salida se mostrará en la misma
posición.
ES-8 STARTER 5000 Medidor de pH de banco
Volver a la interfaz principal
Volver al menú anterior
Entrar, confirmar o aplicar la configuración actual, terminar la configuración
Editar
Retroceder página
Avanzar página
Eliminar
Incrementar (luz)
Reducir (luz)
Pantalla de gráfico (para los datos de medición o calibración)
Guardar
Salida a dispositivo externo (unidad USB, impresora)
Teclado completo
Siguiente punto de calibración
Punto de calibración actual nuevamente
Fin de calibración
STARTER 5000 Medidor de pH de banco ES-9
3.4 Directorio del menú
Los menús del primer y segundo directorio para el STARTER 5000:
Configuración del
sistema
Fecha
Hora
Modo Guardar
Salida
Brillo de la pantalla
Tasa de baudios
Protección de pantalla
Restaurar configuración de fábrica
Información del instrumento
Contraseña de servicio
Configuración de
medición
Resolución
Entrada de temperatura MTC
Modo pH/mV/Rel.mV
Medición continua
Modo Punto final
Configuración de identificación
Modo Medición GLP
Configuración de
calibración
Grupo de amortiguador
Recordatorio de calibración
Datos
Datos de medición
Datos de calibración
ES-10 STARTER 5000 Medidor de pH de banco
4 CALIBRACIÓN, MEDICIÓN E IMPRESIÓN
La pantalla del medidor se apagará automáticamente de acuerdo con la hora configurada en el
menú System Setting (Configuración del sistema) ►Screen Protection (Protección de
pantalla) ►LCD Auto-off. (Apagado autotico de LCD) Para volver a encender la pantalla,
toque la pantalla (vea 3.1)
El proceso de medición de pH deberá seguir este procedimiento:
a) Enjuagar el electrodo de pH y conectarlo al medidor
b) Preparar el amortiguador y calibrar el electrodo de pH
c) Preparar la muestra
d) Realizar la medición de pH de la muestra
e) Registrar los datos de la medición
f) Enjuagar el electrodo de pH y almacenarlo
Preparación del electrodo del pH: el electrodo de pH deberá enjuagarse con agua pura antes y
después de usarse. Revise si el electrodo presenta algún daño sico. (Tenga cuidado con el
bulbo de vidrio.)
El electrodo de pH debe almacenarse en el frasco de almacenamiento. Después de que el
electrodo de pH se haya puesto en la muestra o solución amortiguadora, agite varios segundos y
luego espere de 30 a 60 segundos para que la señal sea estable, y entonces presione el botón
para operar (calibración o medición).
ADVERTENCIA No opere el equipo en ambientes peligrosos. El equipo no tiene
protección contra explosión.
PRECAUCIÓN Cuando use químicos y solventes, siga las instrucciones del
fabricante del químico y las reglas de seguridad generales de laboratorio.
4.1 Calibración
4.1.1 Grupo de amortiguador
El STARTER 5000 puede realizar calibraciones de 1 a 5 puntos para grupos predeterminados de
amortiguador. Para el grupo de amortiguador autodefinido, puede realizar calibración de 1 a 9
puntos.
Antes de la calibración, asegúrese de que el grupo de amortiguador actual es el grupo deseado.
Cuando use un nuevo electrodo de pH, deberá realizarse una calibración antes de cualquier
medición. Si un electrodo de pH no se usa durante varios as, deberá realizarse una calibración
antes del siguiente uso.
La calibración debe mostrar el valor correcto de pH cuando el medidor reciba la señal del valor de
mV del electrodo de pH.
Pendiente: el coeficiente lineal entre mV y pH de acuerdo con el valor teórico (por
ejemplo, -59.16 mV/pH a 25 °C significa una pendiente del 100%);
Desfase: el valor mV cuando el valor del pH es 7.00. (El valor teórico es 0 mV.)
STARTER 5000 Medidor de pH de banco ES-11
El STARTER 5000 presentará un ícono de condición de electrodo para mostrar el desempeño del
electrodo de pH de acuerdo con los datos de calibración. Normalmente, si vemos un ícono de
cara triste después de una calibración adecuada, es posible que necesite reemplazar el electrodo
de pH por uno nuevo.
El STARTER 5000 se corrige automáticamente para la dependencia de temperatura de los
valores de pH del amortiguador. Por ejemplo, para el grupo de amortiguador estándar de Estados
Unidos "1.68, 4.01, 7.00, 10.01", consulte la siguiente tabla:
Por ejemplo, con el amortiguador de calibración 10.01 a 15 °C, el resultado de calibración deberá
ser pH 10.12.
4.1.2 Calibración
Después de enjuagar el electrodo de pH, coloque el electrodo en el primer amortiguador, agite
durante 10 segundos y luego presione para iniciar la calibración. Cuando está en
calibración, el ícono de señal destella. Cuando se ha alcanzado el punto final, aparece
mientras desaparece y la calibración del punto actual es finalizada. Ahora hay 3
opciones:
Salir de la calibración y volver a la interfaz principal
Terminar la calibración del punto actual e ir a la interfaz del resultado de calibración
Hacer la calibración del punto siguiente
Nota: durante el proceso de calibración, el ST5000 siempre es en Auto Endpoint Mode (Modo
de punto final automático).
Nota: se recomienda usar un electrodo de pH 3 en 1 que incluya el sensor de temperatura o un
electrodo de temperatura aparte. Cuando use el modo MTC, ingrese el valor de temperatura
correcto y mantenga todas las soluciones amortiguadoras y de muestra a la temperatura
establecida.
5 °C
1.67
4.01
7.09
10.25
10 °C
1.67
4.00
7.06
10.18
15 °C
1.67
4.00
7.04
10.12
20 °C
1.68
4.00
7.02
10.06
25 °C
1.68
4.01
7.00
10.01
30 °C
1.68
4.01
6.99
9.97
35 °C
1.69
4.02
6.98
9.93
40 °C
1.69
4.03
6.97
9.89
45 °C
1.70
4.05
6.97
9.86
50 °C
1.71
4.06
6.96
9.83
ES-12 STARTER 5000 Medidor de pH de banco
4.2 Medición de muestras
4.2.1 Medición del pH
Asegúrese de que el electrodo se haya calibrado antes de la medición. Después de enjuagar,
Coloque el electrodo de pH en la muestra.
Presione para comenzar la medición de pH; cuando se haya alcanzado el punto final,
la medición actual es terminada.
Nota: si la señal no es estable todo el tiempo, se recomienda usar el modo de punto final manual
para evitar la repetibilidad errónea.
4.2.2 Medición de mV y Rel.mV
Presione para cambiar al modo o ; siga el mismo procedimiento
que para la medición del pH para la medición de mV o Rel.mV.
Si el medidor está conectado con un electrodo ORP (potencial de oxidación reducción), el
resultado de mV es el valor del valor “potencial de oxidación-reducción”, también llamado valor
Redox.
4.3 Medición de temperatura
Para mayor precisión, recomendamos usar un electrodo de temperatura integrado o separado.
Si se usa un electrodo de temperatura, se muestran ATC y la temperatura de la muestra.
Si el medidor no detecta un electrodo de temperatura, automáticamente cambia al modo de
compensación de temperatura manual y aparece MTC. Se debe establecer la temperatura
MTC.
Nota: el STARTER 5000 acepta sensor de temperatura NTC de 30 . Para ATC (compensación
de temperatura automática) o MTC (compensación de temperatura manual), la compensación de
temperatura solo se corrige para el cambio en la salida del electrodo, no para el cambio en la
solución real. Eso significa que el medidor corrige la señal (mV) de los electrodos de pH para
obtener un valor de pH más preciso de acuerdo con la temperatura real.
STARTER 5000 Medidor de pH de banco ES-13
4.4 Impresión
El STARTER 5000 puede imprimir los datos de medición o calibración a través del puerto RS232.
Después de que la impresora (por ejemplo, OHAUS SF-40A) se conecte al medidor
correctamente, realice las siguientes configuraciones (vea 5.1):
ordene el cable de conexión para ST5000 (N/P 30059316) para conectar SF40A y ST5000.
System Setting (Configuración del sistema) Output (Salida) , seleccione Printer
(Impresora)
System Setting (Configuración del sistema) Baud Rate (Tasa de baudios), seleccione
9600.
El formato para la impresión de la medición es:
Fecha Hora Usuario Id. de la muestra
6-13-2015 14:20:22 Usuario01 Muestra02
Id. sensor Modo punto final MTC/ATC
Sensor02 Manual ATC
Temperatura Resultado
25.3 ºC 6.999pH
El formato para la impresión de calibración es:
Datos de calibración
Id. de sensor mero de serie
Sensor02 14060012
Fecha cal. Hora cal. ATC/MTC
01-01-2015 14:20:33 ATC
Amortig. mV Desfase Pendiente
4.00pH 176.94mV
7.00pH 0.05mV 0.05mV 98.22%
Cuando se realizan mediciones continuas, el formato para imprimir los datos de calibración es:
Tiempo Medición Registro
Id. de sensor Número de serie
Sensor02 14060012
Fecha Hora Temperatura Medición
01-01-2015 14:20:25 23.5 ºC 6.998pH
Fecha Hora Temperatura Medición
01-01-2015 14:20:30 23.5 ºC 6.999pH
Fecha Hora Temperatura Medición
01-01-2015 14:20:35 23.5 ºC 7.000pH
ES-14 STARTER 5000 Medidor de pH de banco
5 CONFIGURACIÓN
5.1 Configuración del sistema
La configuración del sistema se usa para establecer parámetros del medidor; la información del
medidor también podría encontrarse en este. Vea la sección 3.4 para referencia.
5.1.1 System setting (Configuracn del sistema) Save Mode (Modo
Guardar)
El modo Guardar tiene 2 opciones, uno es “Auto Save Endpoint Reading” (Guardar
automáticamente lectura del punto final) y el otro es “Manual Save Endpoint Reading” (Guardar
manualmente la lectura del punto final). “Auto Save Endpoint Reading” significa que cuando la
medición alcanza el punto final (punto final automático, manual u hora), el medidor guardará
automáticamente la lectura del punto final.
El ícono Guardar cambiará de negro a gris claro ; una lectura del punto final
solo puede guardarse una vez.
5.1.2 System Setting (Configuración del sistema) ►Output (Salida)
Cuando se selecciona apagar la conexn, la interfaz principal mostrará el ícono
Guardar , o mostrará el ícono Salida .
5.1.3 System Setting (Configuración del sistema) Restore Factory Setup
(Restaurar configuracn de fábrica)
La restauración de la configuración de fábrica incluye lo siguiente:
1. Apagar el modo GLP
2. Apagar el modo Guardar automáticamente
3. Apagar la medición continua
4. Apagar la conexión (salida)
5. Apagar el recordatorio de calibración
6. La resolución es: 0.001pH, 0.01mV
7. Modo de medición: pH:
8. Se eliminará el grupo de amortiguador autodefinido
9. Tasa de baudios: 9600
10. Brillo de la pantalla: 70%
11. Apagar la protección de pantalla
12. MTC: 25 ºC
13. Punto final: Autotico
14. Punto final de tiempo: 10 segundos
15. Eliminar todos los datos de calibración
5.1.4 System Setting (Configuración del sistema) Service Password
(Contraseña de servicio)
No se usa.
STARTER 5000 Medidor de pH de banco ES-15
5.2 Configuración de medición
5.2.1 Measurement Setting (Configuración de medición) Endpoint Mode
(Modo de punto final)
Punto final autotico
En este modo, el medidor alcanzará el punto final y bloqueará la lectura automáticamente de
acuerdo con los siguientes criterios de estabilidad:
Resolucn 0.1pH: la señal de la entrada del sensor puede no cambiar más de 0.3 mV en 6 segundos.
Resolucn 0.01pH: la señal de la entrada del sensor puede no cambiars de 0.1 mV en 6 segundos.
Resolucn 0.001pH: la señal de la entrada del sensor puede no cambiar s de 0.03 mV en 6
segundos.
Punto final de tiempo
En este modo, el usuario necesita ingresar el tiempo del punto final, por ejemplo 15 segundos.
Esto significa que después de iniciar la medición durante 15 segundos, el medidor alcanzará el
punto final.
Punto final manual
En este modo, el usuario necesita decidir cuándo es estable la señal y luego presionar el ícono
para detener para alcanzar el punto final manual.
5.2.2 Measurement Setting (Configuración de medición) ►GLP
Measurement Mode (Modo de medición GLP)
El modo GLP requiere que cada medición tenga la identificación del usuario y la identificación de
la muestra. Si no hay identificación del usuario e identificación de la muestra, el medidor mostrará
la interfaz para permitir que el usuario ingrese los datos requeridos.
5.2.3 Measurement Setting (Configuración de medición) ►Continuous
Measure (Medición continua)
Durante la medición continua, el usuario necesita ingresar el intervalo de tiempo para obtener
cada resultado de lectura. A continuación se muestra un Gráfico de medición continua.
El eje Y es el valor de pH. El eje x es el número del resultado de la medición.
ES-16 STARTER 5000 Medidor de pH de banco
5.3 Configuración de calibracn
5.3.1 Calibration Setting (Configuración de calibración) ►Buffer Group
(Grupo de amortiguador)
Antes de la calibracn, asegúrese de que el grupo de amortiguador actual es el correcto para sus
soluciones amortiguadoras.
El usuario puede seleccionar entre 8 grupos de amortiguador definidos.
El usuario también puede establecer grupos de amortiguador autodefinidos. La diferencia de 2
puntos de calibración adyacentes deberá ser mayor de 0.99pH.
5.3.2 Calibration Setting (Configuración de calibración) ►Calibration
Reminder (Recordatorio de calibración)
Ingrese a esta configuración para ajustar el tiempo, desde la última calibración, para determinar
después de cuántas horas el medidor debe recordar realizar la calibración nuevamente. El
medidor mostrará el ícono de recordatorio de calibración y destellará para recordar al
usuario realizar la calibración.
5.4 Registro de datos
Toque Datos para entrar en la interfaz del registro de datos, incluyendo datos de medición y
datos de calibración.
Aquí puede eliminar todos los datos de medición. La única forma de eliminar datos de calibración
es restaurando la configuración de fábrica.
Los datos pueden enviarse a una unidad USB (≤ 4 GB).
STARTER 5000 Medidor de pH de banco ES-17
6 MANTENIMIENTO
6.1 Mensaje de error
Cuando el valor ingresado o resultado de la medición esté fuera de rango o los datos de
calibración no sean suficientemente buenos, el medidor mostrará información de error.
Si el error no puede resolverse mediante el mensaje de error mostrado, comuníquese con el
representante de ventas de Ohaus para mayor asistencia.
6.2 Mantenimiento del medidor
Los instrumentos de la serie STARTER 5000 no requieren ningún mantenimiento excepto la
limpieza ocasional con una tela humedecida.
Atención: siga estas precauciones de seguridad cuando limpie el instrumento.
Desconecte el instrumento de la alimentación eléctrica
No abra el instrumento
No permita que los líquidos penetren en el instrumento
Atención: la caja está hecha de acronitrilo butadieno estireno (ABS). Este material es atacado
por algunos solventes orgánicos como el tolueno, el xileno y el metil etil cetona (MEK). Cualquier
derrame deberá limpiarse inmediatamente.
6.3 Mantenimiento del electrodo
Asegúrese de que el electrodo esté lleno con solución de electrólito. Siempre almacene el
electrodo de acuerdo con el manual de instrucciones del electrodo y no permita que se seque.
Si la respuesta del electrodo llega a ser lenta o la pendiente no es suficientemente buena, intente
lo siguiente:
o Remoje el electrodo en HCl 0.1 M durante más de 8 horas.
o Para contaminación con grasa o aceite, desengrase la membrana con lana de algodón
empapada en acetona o solución de jabón.
Después del tratamiento del electrodo, deberá realizarse una nueva calibración. Si la pendiente
del electrodo aún no es buena, el electrodo podría necesitar reemplazarse.
6.4 Información
Toque el ícono de información para entrar en la interfaz de información. El usuario puede
ingresar aquí información relacionada al tocar el área que tiene " " en la parte inferior izquierda,
tal como fecha de compra del sensor y otra.
ES-18 STARTER 5000 Medidor de pH de banco
7 DATOS TÉCNICOS
7.1 Especificaciones
Los datos técnicos son válidos en las siguientes condiciones ambientales:
Para uso exclusivo en interiores
Altitud: hasta 2000 m
Rango de temperatura especificado: 5 °C a 40 °C
Humedad: humedad relativa máxima del 80 % para temperaturas hasta de 30 °C,
disminuyendo linealmente al 50% de humedad relativa a 40 °C
Fluctuaciones de voltaje de la alimentación principal: hasta ± 10% del voltaje nominal
Categoría de instalación II
Grado de contaminación: 2
La operabilidad se garantiza a temperaturas ambiente entre 5 °C a 40 °C
Modelo
STARTER 5000
Rango de medición
-2.000…20.000 pH
–2000.00…2000.00 mV
-30 °C…130 °C
Resolución
0.1/0.01/0.001 pH
1/ 0.1/ 0.01 mV
0.1 °C
mites de error
± 0.002 pH
± 0.03% FS mV
± 0.1 °C
Calibración
1 a 9 puntos
8 grupos de amortiguador predefinidos s 1 grupo de
amortiguador autodefinido
Memoria
1000 mediciones
10*10 datos de calibración
Alimentación de energía
Entrada de 100-240 V ~0.3 A, 9V Adaptador de CA de
1.12 A de salida
Tamaño/peso
Aproximadamente 210 x 170 x 80 mm / 0.65 kg
Dimensión/peso de envío
Aproximadamente 372 x 310 x 185 mm / 2.30 kg
Pantalla
Pantalla táctil de cristalquido con iluminación de fondo
Entrada
BNC, impedancia > 3 x 10e+12 Ω
Cinch, NTC 30
Compensación de temperatura
ATC y MTC
Caja
ABS
STARTER 5000 Medidor de pH de banco ES-19
8 CONFORMIDAD NORMATIVA
Este producto está en conformidad con la Directiva EMC 2004/108/EC y con la
Directiva para Bajo Voltaje 2006/95/EC. La Declaración de conformidad está
disponible en línea en
europe.ohaus.com/europe/en/home/support/compliance.aspx.
De conformidad con la Directiva Europea 2002/96/EC sobre Residuos de
Equipos Eléctricos y Electrónicos (WEEE), este dispositivo no puede
desecharse con la basura doméstica. Esto también es aplicable para países
fuera de la UE, según sus requerimientos específicos. La Directiva para
baterías 2006/66/EC introduce nuevos requisitos a partir de septiembre de 2008
acerca de la remoción de las baterías de equipos de desecho en los países
miembros de la Unión Europea. Para cumplir con esta Directiva, este dispositivo
ha sido diseñado para remoción segura de baterías al final de su vida útil a
través de un centro de tratamiento de desechos. Deseche este producto de
acuerdo con las regulaciones locales en el punto de recolección especificado
para equipos eléctricos y electrónicos. Si tiene alguna pregunta, comuníquese
con la autoridad responsable o con el distribuidor a quien compró este
dispositivo. En caso que este dispositivo sea transferido a otras partes (para
uso privado o profesional), también deberá mencionarse el contenido de esta
regulación. Las instrucciones de desecho en Europa esn disponibles en línea
en europe.ohaus.com/europe/en/home/support/weee.aspx. Gracias por su
contribución a la protección ambiental.
Nota de la FCC: Este equipo ha sido probado y se encontró que cumple con los límites
para un dispositivo digital clase A, consecuente con la Parte 15 de las Reglas de la FCC.
Estos límites están diseñados para proporcionar una protección razonable contra
interferencia dañina cuando el equipo es operado en un ambiente comercial. Este equipo
genera, usa y puede irradiar frecuencias de radio y, si no es instalado y utilizado de
acuerdo con el manual de instrucciones, puede causar interferencia dañina a las
radiocomunicaciones. Es probable que la operación de este equipo en un área residencial
cause interferencia dañina, en cuyo caso se le exigirá al usuario que corrija la interferencia
con gastos a su cargo.
Nota de Industry Canada
Este aparato digital Clase A cumple con la norma canadiense ICES-001.
Certificación ISO 9001
En 1994, Ohaus Corporation de Estados Unidos recibió la certificación ISO 9001 de parte
de Bureau Veritas Quality International (BVQI), lo cual confir que el sistema de gestión
de calidad de Ohaus cumple con los requerimientos de la norma ISO 9001. El 21 de junio
de 2012, Ohaus Corporation de Estados Unidos recibió la recertificación de la norma ISO
9001:2008.
ES-20 STARTER 5000 Medidor de pH de banco
9 GRUPOS DE AMORTIGUADOR
El STARTER 5000 corrige automáticamente la dependencia de temperatura del valor de pH del
grupo de amortiguador dado en las siguientes tablas. Seleccione el grupo de amortiguador que
desea antes de la calibración.
JJG119 (Ref. 25 °C, 0.01pH)
Temp. °C 1.68 4.00 6.86 9.18 12.46
5 1.67 4.00 6.95 9.39 13.21
10 1.67 4.00 6.92 9.33 13.01
15 1.67 4.00 6.90 9.28 12.82
20 1.68 4.00 6.88 9.23 12.64
25 1.68 4.00 6.86 9.18 12.46
30 1.68 4.01 6.85 9.14 12.29
35 1.69 4.02 6.84 9.11 12.13
40 1.69 4.03 6.84 9.07 11.98
45 1.70 4.04 6.83 9.04 11.83
50 1.71 4.06 6.83 9.02 11.70
Estándar de los EE. UU. (Ref. 25 °C)
Temp. °C 1.68 4.01 7.00 10.01
5 1.67 4.01 7.09 10.25
10 1.67 4.00 7.06 10.18
15 1.67 4.00 7.04 10.12
20 1.68 4.00 7.02 10.06
25 1.68 4.01 7.00 10.01
30 1.68 4.01 6.99 9.97
35 1.69 4.02 6.98 9.93
40 1.69 4.03 6.97 9.89
45 1.70 4.05 6.97 9.86
50 1.71 4.06 6.96 9.83
Estándar de Europa (Ref. 25 °C)
Temp. °C 2.00 4.01 7.00 9.21 11.00
5 2.02 4.01 7.09 9.45 11.72
10 2.01 4.00 7.06 9.38 11.54
15 2.00 4.00 7.04 9.32 11.36
20 2.00 4.00 7.02 9.26 11.18
25 2.00 4.01 7.00 9.21 11.00
30 1.99 4.01 6.99 9.16 10.82
35 1.99 4.02 6.98 9.11 10.64
40 1.98 4.03 6.97 9.06 10.46
45 1.98 4.04 6.97 9.03 10.28
50 1.98 4.06 6.97 8.99 10.10
STARTER 5000 Medidor de pH de banco ES-21
MERCK (Ref. 20 °C)
Temp. °C 2.00 4.00 7.00 9.00 12.00
5 2.01 4.04 7.07 9.16 12.41
10 2.01 4.02 7.05 9.11 12.26
15 2.00 4.01 7.02 9.05 12.10
20 2.00 4.00 7.00 9.00 12.00
25 2.00 4.01 6.98 8.95 11.88
30 2.00 4.01 6.98 8.91 11.72
35 2.00 4.01 6.96 8.88 11.67
40 2.00 4.01 6.95 8.85 11.54
45 2.00 4.01 6.95 8.82 11.44
50 2.00 4.00 6.95 8.79 11.33
DIN (19267) (Ref. 25 °C)
Temp. °C 1.09 4.65 6.79 9.23 12.75
5 1.08 4.67 6.87 9.43 13.63
10 1.09 4.66 6.84 9.37 13.37
15 1.09 4.66 6.82 9.32 13.16
20 1.09 4.65 6.80 9.27 12.96
25 1.09 4.65 6.79 9.23 12.75
30 1.10 4.65 6.78 9.18 12.61
35 1.10 4.65 6.77 9.13 12.45
40 1.10 4.66 6.76 9.09 12.29
45 1.10 4.67 6.76 9.04 12.09
50 1.11 4.68 6.76 9.00 11.98
JIS Z 8802 (Ref. 25 °C)
Temp. °C 1.679 4.008 6.865 9.180
5 1.668 3.999 6.951 9.395
10 1.670 3.998 6.923 9.332
15 1.672 3.999 6.900 9.276
20 1.675 4.002 6.881 9.225
25 1.679 4.008 6.865 9.180
30 1.683 4.015 6.853 9.139
35 1.688 4.024 6.844 9.102
40 1.694 4.035 6.838 9.068
45 1.700 4.047 6.834 9.038
50 1.707 4.060 6.833 9.011
DIN(19266) / NIST (Ref. 25 °C)
Temp. °C 1.680 4.008 6.865 9.183 12.454
5 1.668 4.004 6.950 9.392 13.207
10 1.670 4.001 6.922 9.331 13.003
15 1.672 4.001 6.900 9.277 12.810
20 1.676 4.003 6.880 9.228 12.627
25 1.680 4.008 6.865 9.183 12.454
30 1.685 4.015 6.853 9.144 12.289
35 1.691 4.026 6.845 9.110 12.133
40 1.697 4.036 6.837 9.076 11.984
45 1.704 4.049 6.834 9.046 11.841
50 1.712 4.064 6.833 9.018 11.705
ES-22 STARTER 5000 Medidor de pH de banco
JJG119 (Ref. 25 °C, 0.001pH)
Temp. °C 1.680 4.003 6.864 9.182 12460
5 1.669 3.999 6.949 9.391 13.210
10 1.671 3.996 6.921 9.330 13.011
15 1.673 3.996 6.898 9.276 12.820
20 1.676 3.998 6.879 9.226 12.637
25 1.680 4.003 6.864 9.182 12.460
30 1.684 4.010 6.852 9.142 12.292
35 1.688 4.019 6.844 9.105 12.130
40 1.694 4.029 6.838 9.072 11.975
45 1.700 4.042 6.834 9.042 11.828
50 1.706 4.055 6.833 9.015 11.697
STARTER 5000 Medidor de pH de banco ES-23
GARANTÍA LIMITADA
Los productos de Ohaus están garantizados contra defectos de materiales y fabricación desde la
fecha de entrega y durante el tiempo que dure la garantía. Durante el periodo de garantía Ohaus
reparará, o, a su discreción, reemplazará cualquier componente(s) que se compruebe que está
defectuoso, sin costo alguno, con la condición de que el producto sea devuelto, flete pagado, a
Ohaus.
Esta garantía no es lida si el producto ha sufrido daños por accidente o mal uso, ha sido
expuesto a materiales radioactivos o corrosivos, presenta materiales extraños penetrando al
interior del producto, o si ha sido objeto de modificaciones por parte de personas que no sean de
Ohaus. En el lugar de una tarjeta de registro de garantía devuelta debidamente, el periodo de
garantía comenzará en la fecha de envío al distribuidor autorizado. Ohaus Corporation no otorga
ninguna otra garantía expresa o implícita. Ohaus Corporation no se hace responsable de
cualquier daño resultante.
Puesto que la legislación sobre garantías difiere de un estado a otro así como de un país a otro,
comuníquese con su distribuidor local de Ohaus para mayor información.
TABLE DES MATIÈRES
1 INTRODUCTION................................................................ 1
1.1 finition des symboles et des indicateurs d'avertissements ... 1
1.2 Précautions de sécurité .................................................................. 2
2 INSTALLATION ................................................................. 3
2.1 Contenu de l’emballage ................................................................... 3
2.2 Installation du porte-électrode autonome ..................................... 4
2.3 Installation de l'adaptateur d'alimentation .................................... 4
2.4 Connexion des électrodes de pH ................................................... 5
3 ICÔNES ET AFFICHAGE .................................................. 5
3.1 Mise sous tension de l’appareil ...................................................... 5
3.2 Mise hors tension de l'appareil ....................................................... 6
3.3 Icônes ................................................................................................ 6
3.4 pertoire du menu ......................................................................... 9
4 ÉTALONNAGE, MESURE ET IMPRESSION .................. 10
4.1 Étalonnage ...................................................................................... 10
4.1.1 Groupe de tampons ......................................................................... 10
4.1.2 alisation d'un étalonnage .............................................................. 11
4.2 Mesure d'un échantillon ................................................................ 12
4.2.1 Mesure du pH .................................................................................. 12
4.2.2 Mesure en mV et Rel.mV ................................................................. 12
4.3 Mesure de température ................................................................. 12
4.4 Impression ...................................................................................... 13
5 CONFIGURATION ........................................................... 14
5.1 Paramétrage système .................................................................... 14
5.1.1 Paramétrage systèmeMode Enregistrement................................ 14
5.1.2 Paramétrage systèmeSortie ......................................................... 14
5.1.3 Paramétrage systèmeRestauration des paramètres par défaut .. 14
5.1.4 Paramétrage système Mot de passe de service .......................... 14
5.2 Paramétrage des mesures ............................................................ 15
5.2.1 Paramétrage des mesures Mode de résultat final ....................... 15
5.2.2 Paramétrage des mesures Mode GLP de mesure ...................... 15
5.2.3 Paramétrage des mesures Mesure en continu ............................ 15
5.3 Paramétrage de l'étalonnage ........................................................ 16
5.3.1 Paramétrage de l'étalonnage Groupe de tampons ...................... 16
5.3.2 Paramétrage de l'étalonnage Rappel de l'étalonnage ................. 16
5.4 Journal des données ..................................................................... 16
6 MAINTENANCE............................................................... 17
6.1 Message d’erreur ........................................................................... 17
6.2 Maintenance de l'appareil de mesure .......................................... 17
6.3 Maintenance de l'électrode ........................................................... 17
6.4 Informations ................................................................................... 17
7 DONNÉES TECHNIQUES ............................................... 18
7.1 Spécifications ................................................................................. 18
8 CONFORMITÉ ................................................................. 19
9 GROUPES DE TAMPONS .............................................. 20
PH-mètre de laboratoire STARTER 5000 FR-1
1 INTRODUCTION
Merci d'avoir choisi ce produit OHAUS.
Lire le manuel complètement avant d'utiliser le pH-mètre de laboratoire STARTER 5000 afin
d'éviter une exploitation inadéquate.
Le STARTER 5000 est un pH-mètre de laboratoire (0,001) avec un écran ACL tactile couleur. Il s'agit
d'une nouvelle ration d'appareils intuitifs et faciles d'utilisation permettant des mesures
fiables.
Suivent ici quelques-unes des nombreuses fonctions existantes :
Interface utilisateur graphique bilingue
Indicateur temporel de résultats automatiques et manuels pouvant aider les utilisateurs à
figer un résultat stable afin de répondre aux différents impératifs de la clientèle
8 groupes de tampons (US, EU, NIST et JJG) avec un tampon auto-défini
Protection IP54 contre l'eau et la poussière
1.1 finition des symboles et des indicateurs d'avertissements
Les remarques de curité sont signalées par des mots indicateurs et par des symboles
d'avertissement. Elles présentent les problèmes et les avertissements de sécurité. L'inobservation
des remarques de curité peut conduire à des accidents, à l'endommagement de l'instrument, à
des dysfonctionnements et à des résultats erronés.
Mots indicateurs
AVERTISSEMENT Pour une situation dangereuse avec un risque moyen pouvant
être à l'origine d'un accident ou d'un dés, s'il n'est pas évité.
PRÉCAUTIONS Pour une situation dangereuse avec un faible risque pouvant être
à l'origine de dommages au dispositif ou aux biens, d'une perte des
données, ou d'un accident, si elle n'est pas évitée.
Attention Pour des informations importantes sur ce produit
Remarque Pour des informations utiles sur le produit
Symboles d'avertissement
Danger géral Danger de choc électrique
Risque d'explosion Danger, produits corrosifs
Tension alternative Courant continu
FR-2 PH-mètre de laboratoire STARTER 5000
1.2 Précautions de sécurité
ATTENTION Lire tous les avertissements de sécurité avant d’effectuer l’installation,
les connexions ou les réparations de cet équipement. Si ces avertissements ne sont
pas respectés, des accidents et/ou des dommages matériels peuvent en sulter.
Conserver toutes les instructions pour rérence ultérieure.
rifier que la plage de tension d'entrée de l'adaptateur CA et le type de prise sont
compatibles avec l'alimentation alternative locale.
S'assurer que le cordon d'alimentation n’est pas un obstacle potentiel ou ne présente aucun
danger de chute.
Utiliser l'équipement seulement dans des lieux secs.
cher immédiatement tout déversement de liquide. L'instrument n'est pas étanche.
Utiliser uniquement les accessoires et les périphériques approus.
Faire fonctionner lquipement uniquement dans les conditions ambiantes spécifiées dans ces
instructions.
connecter l'équipement de l'alimentation électrique pour le nettoyer.
Ne pas utiliser lquipement dans des environnements dangereux ou instables.
La maintenance doit être exclusivement exécutée par un personnel agréé.
Si l'équipement est utili de façon non spécifiée par le fabricant, laprotection assurée par l'éq
uipement peut être compromise.
AVERTISSEMENT Ne pas utiliser cet instrument dans un environnement pouvant
présenter des dangers d'explosion. L'enceinte de l'instrument n'est pas étanche aux
gaz (danger d'explosion en présence d'étincelles, corrosion provoquée par
l'introduction des gaz).
ATTENTION Lors de l'utilisation de produits chimiques et de solvants, se conformer
aux instructions du fabricant et aux règles rales de sécurité du laboratoire.
PH-mètre de laboratoire STARTER 5000 FR-3
2 INSTALLATION
baller soigneusement la boîte.
2.1 Contenu de l’emballage
Le modèle ST5000-B (conditionnement de base) contient les articles suivants :
En complément au contenu du ST500-B, l'emballage du ST5000-F contient aussi ce qui suit :
La poudre pour tampon de pH doit être dissoute dans 250 ml d'eau pure ou sionisée dans une
fiole jaugée. Il est aussi possible de commander auprès d'Ohaus des solutions de tampons en
bouteille.
Contacter Ohaus pour de plus amples tails sur les électrodes, les solutions, les accessoires et
les pièces de rechange.
Modèle
Description
rence
ST350
Électrode de pH à remplissage en verre 3 en 1
30129354
ST310
Électrode de pH à remplissage en plastique 3 en 1
83033965
ST210
Électrode de pH à remplissage en plastique 2 en 1
83033966
ST320
Électrode de pH sous forme de gel en plastique 3 en 1 (nul
besoin de remplissage)
83033967
ST260
Électrode de pH pour double pont salin en verre 2 en 1
30129357
ST230
Électrode de pH en verre pour échantillons troubles 2 en 1
83033968
STSURF
Électrode de pH pour surface plane 2 en 1
30129470
STMICRO5
Électrode de pH pour microéchantillons 2 en 1
30087566
STMICRO8
Électrode de pH pour microéchantillons en verre 2 en 1
30087569
STTEMP30
Électrode de température
83033970
STORP1
Électrode ORP en plastique pour gel
30038555
STORP2
Électrode réutilisable ORP en verre
30038553
ST5000-B
Unités
pH-mètreST5000
1
Couvercle en utilisation ST5000
1
Porte-électrode autonome
1 ensemble
Clé USB 4Go
1
Poudre pour tampon de pH (4,01 ; 7,00 ; 10,00)
1 ensemble
Alimentation
1 ensemble
Électrode de pH à remplissage en verre ST350 3 en 1
1
Poudre pour tampon (4,01 ; 7,00 ; 10,0)
83033971
Tampon de pH 1,68 250 ml
30100424
Tampon de pH 4,01 250ml
30100425
Tampon de pH 7,00 250ml
30100427
Tampon de pH 10,01 250ml
30100429
Tampon de pH 12,45 250ml
30100440
Électrolyte de rérence pour électrode de pH
30059255
Solution de protection pour électrode de pH
30059256
Porte-électrode autonome
30058733
Couvercle en utilisation ST5000
30129897
Adaptateur deble SP RS232 F-M SF40A (pour ST5000)
30059316
FR-4 PH-mètre de laboratoire STARTER 5000
2.2 Installation du porte-électrode autonome
Installer le bras de l'électrode sur la base :
Après avoir glé le bouton de tension jusqu'à un certain niveau, il est possible de placer
le bras supérieur vers le haut et vers le bas. En plaçant l'électrode de pH sur le bras et en
alisant un étalonnage ou une mesure, il est possible de placer la bouteille de stockage de
l'électrode de pH dans l'orifice de la base pour un rangement aisé:
2.3 Installation de l'adaptateur d'alimentation
Insérer la prise de l'adaptateur dans la fente de l'adaptateur d'alimentation.
ATTENTION N'utiliser qu'une prise d'adaptateur qui est compatible avec la prise
secteur. S'assurer que la prise de l'adaptateur est bien enfone dans la fente de
l'adaptateur d'alimentation avant de la connecter à la prise secteur.
PH-mètre de laboratoire STARTER 5000 FR-5
2.4 Connexion des électrodes de pH
Il existe 2 supports d'électrodes : support pH (BNC), support TEMP (Cinch).
Pour l'électrode ST350 3 en 1, la connecter aux prises BNC et Cinch. Pour les électrodes de pH
2 en 1 (par exemple, ST230), les connecter seulement à la prise BNC.
3 ICÔNES ET AFFICHAGE
3.1 Mise sous tension de l’appareil
Le STARTER 5000 est un pH-tre de laboratoire 0,001 en couleurs et à écran tactile. Il ne
comporte aucun bouton.
Après l'avoir connecté à l'alimentation, le pH-mètre se mettra sous tension ; si la protection de
l'écran est activée, appuyer sur l'écran pour l'activer.
Remarque : Si le client n'utilise pas l'appareil pendant plus de 6 heures, ce dernier se mettra
automatiquement hors tension. Il n'est plus possible d'appuyer sur l'écran pour le activer. Il doit
être reconnecté à l'alimentation pour le remettre sous tension.
lectionner la langue et appuyer ensuite sur pour accéder à l'interface principale.
Interface principale
FR-6 PH-mètre de laboratoire STARTER 5000
3.2 Mise hors tension de l'appareil
Pour mettre l'appareil hors tension, appuyer sur le bouton de mise hors tension en partie
supérieure gauche de l'interface principale pour accéder à l'interface de mise hors tension.
Appuyer ensuite sur pour mettre l'appareil hors tension, ou sur pour
revenir à l'interface principale.
Mise hors tension de l'interface
3.3 Icônes
Mode de mesure du pH. Passer entre les modes mV et Rel.mV ou
Icônes de résultat final, de résultat final automatique, de résultat final manuel et de
sultat final temporel.
Avec un résultat final manuel, si le signal est stable, l'appareil affichera l'icône d'un
signal stable. Appuyer sur l'icône d'arrêt pour atteindre le résultat final manuel.
PH-mètre de laboratoire STARTER 5000 FR-7
Zone de touche des informations, des mesures et de l'étalonnage
Cette zone d'informations sert aussi de raccourci pour entrer sur l'interface de paramétrage :
«sultat final :
»
Appuyer sur cette zone pour entrer sur l'interface afin definir le mode de
sultat final. Les modes disponibles du résultat final sont Automatique,
Manuel et Temporisé. Le mode desultat final actuel est affiché.
"Capteur ID :"
Appuyer sur cette zone pour entrer sur l'interface et lectionner ou modifier
l'ID du capteur. L'ID du capteur actuel est affichée.
"Condition
capteur :"
Toucher cette zone pour entrer sur l'interface des données d'étalonnage du
capteur. L'icône des performances du capteur actuel (établies par rapport
aux données d'étalonnage) est affichée.
Il existe 3 icônes de conditions du capteur :
Pente : plus de 95 %
et décalage : ± (0-15) mV
Bonne condition de l'électrode
Pente : 90-95 %
ou décalage : ± (15-35) mV
Condition acceptable de
l'électrode
Pente : moins de 90 %
ou décalage : ± (35-60) mV
La condition de llectrode
n'est pas bonne ou nécessite
un nettoyage
Icône de rappel de l'étalonnage. Définir l’heure à laquelle l'appareil doit signaler
que la durée depuis le dernier étalonnage est écoulée. Lorsque la durée est
atteinte, cette ine s'affichera et clignotera pour rappeler qu'il est temps de
aliser un étalonnage.
Tampons :
Cette zone affiche le groupe actuel de tampons, appuyer sur cette zone pour
entrer sur l'interface de sélection du groupe de tampons.
Icône de démarrage. Appuyer sur cette icône et l'appareil commencera la
mesure, l'icône de signal clignotera. La zone de l'icône de démarrage
affichera l'icône d'arrêt
Icône d'étalonnage. Appuyer sur cette icône pour démarrer l'étalonnage.
Icône d'enregistrement. Après avoir enregistré le résultat de la mesure en cours
(résultat final atteint), la couleur de cette icône passera de noir à gris clair ,
ce qui indique que la mesure a dé été enregistrée et ne peut l'être une
deuxième fois.
Icône de sortie. Diriger les données de sortie vers un dispositif externe tel qu'une
imprimante.
Si « Arrêt connexion » est choisi dans Paramétrage système ► Sortie, l'icône
d'enregistrement s'affichera. Sinon, l'icône de sortie s'affichera sur la
me position.
FR-8 PH-mètre de laboratoire STARTER 5000
Retour à l'interface principale
Retour au menu prédent
Entrer, confirmer ou appliquer la configuration actuelle et terminer le paramétrage
Modification
Page précédente
Page suivante
Suppression
Augmentation (éclairage)
Diminution (éclairage)
Affichage graphique (pour les données de mesurage ou d'étalonnage)
Enregistrement
Sortie vers un dispositif externe (clé USB, imprimante)
Clavier complet
Point d'étalonnage suivant
Point d'étalonnage en cours
Fin de l'étalonnage
PH-mètre de laboratoire STARTER 5000 FR-9
3.4 pertoire du menu
Le menu du premier et du second répertoire du STARTER 5000 :
Système
Paramètre
Date
Heure
Mode d'enregistrement
Sortie
Luminosité de l'écran
bit en bauds
Protection de l’écran
cupération des paratres usine
Informations sur l'instrument
Mot passe maintenance
Mesure
Paramètre
solution
Température MTC Saisie
Mode pH/mV/Rel.mV
Mesure en continu
Mode de résultat final
Paramétrage ID
Mode de mesure GLP
Étalonnage
Paramètre
Groupe de tampons
Rappel de l'étalonnage
Données
Données de mesure
Données d'étalonnage
FR-10 PH-mètre de laboratoire STARTER 5000
4 ÉTALONNAGE, MESURE ET IMPRESSION
L'affichage de l'appareil s'éteindra automatiquement conforment à la durée définie dans le
menu Paramétrage système ►Protection de l'écran ► Arrêt automatique ACL. Pour
remettre l'affichage sous tension, appuyer sur l'écran. (Voir 3.1)
Le processus complet de mesure du pH doit suivre la procédure :
a) Rincer l'électrode de pH et connecter à l'appareil
b) Préparer le tampon et étalonner l'électrode de pH
c) Préparer l'échantillon
d) aliser la mesure du pH de l'échantillon
e) Enregistrer les données de mesure
f) Rincer l'électrode de pH et ranger
Préparer l'électrode de pH : elle doit être rincée avec de l'eau pure avant et après utilisation.
rifier si l'électrode est endommagée. (Prendre garde à l'ampoule en verre)
L'électrode de pH doit être rangée dans une bouteille de stockage. Après avoir plongé l'électrode
de pH dans l'échantillon ou dans la solution tampon, l'utilisateur doit homogéiser pendant
plusieurs secondes et attendre 30 à 60 secondes que le signal soit stable ; appuyer ensuite sur le
bouton de fonctionnement (étalonnage ou mesure).
AVERTISSEMENT Ne pas utiliser l'équipement dans des environnements dangereux.
L'équipement n'est pas proté contre les explosions.
ATTENTION Avec des produits chimiques et des solvants, se conformer aux
instructions du fabricant desdits produits et aux règles gérales de sécurité du
laboratoire.
4.1 Étalonnage
4.1.1 Groupe de tampons
Le STARTER 5000 peut effectuer des étalonnages 1 et 5 points pour des groupes de tampons
préfinis. Pour le groupe de tampons auto-définis, des étalonnages 1 à 9 points peuvent être
alisés.
Avant l'étalonnage, s'assurer que le groupe actuel de tampons est bien celuisiré.
Lors de l'utilisation d'une nouvelle électrode de pH, un étalonnage doit être alisé avant
d'effectuer des mesures. Si une électrode de pH reste inutilisée pendant quelques jours, un
étalonnage doit être réalisé avant sa prochaine utilisation.
L'étalonnage doit afficher la valeur correcte du pH lorsque l'appareil reçoit un signal en mV
provenant de l'électrode de pH.
Pente : le coefficient linéaire entre mV et pH conformément à la valeur torique (par
exemple, -59,16mV/pH à 25 °C signifie une pente de 100 %),
calage : la valeur en mV lorsque la valeur du pH est égale à 7,00. (La valeur
théorique est de 0 mV)
PH-mètre de laboratoire STARTER 5000 FR-11
Le STARTER 5000 fournira une icône d'état de l'électrode afin de présenter les performances de
l'électrode de pH par rapport aux données d'étalonnage. Normalement, si une icône à « visage
triste » apparaît après un étalonnage adéquat, cela signifie qu'il faut remplacer l'électrode de pH
par une nouvelle.
Le STARTER 5000 corrige automatiquement la pendance à la température des valeurs de pH
du tampon. Par exemple, pour le groupe de tampons étalon US « 1,68, 4,01, 7,00, 10,01 »,
consulter le tableau suivant :
Par exemple, avec un tampon d'étalonnage 10,01 à 15 °C, le sultat de l'étalonnage doit être
10,12 pH.
4.1.2 alisation d'un étalonnage
Après avoir rincé l'électrode de pH, plonger celle-ci dans le premier tampon, remuer pendant 10
secondes et appuyer ensuite sur pour marrer l'étalonnage. Pendant l'étalonnage, l'icône
du signal clignote. Lorsque le résultat final a é atteint, apparaît alors que
disparaît, et le point d'étalonnage en cours est terminé. Il existe maintenant 3 options :
Sortir de l'étalonnage et revenir à l'interface principale
Terminer le point d'étalonnage en cours, se rendre sur l'interface des résultats de
l'étalonnage
Passer au point suivant d'étalonnage
Remarque : Pendant le processus d'étalonnage, le ST5000 est toujours en Mode sultat Final
Automatique.
Remarque : L'utilisation d'une électrode de pH 3 en 1 avec un capteur de température intégré ou
une électrode de température parée est recommandée. Avec une utilisation en mode MTC,
entrer la valeur correcte de température et conserver tous les tampons et solutions d'échantillons
à la température définie.
5 °C
1,67
4,01
7,09
10,25
10 °C
1,67
4,00
7,06
10,18
15 °C
1,67
4,00
7,04
10,12
20 °C
1,68
4,00
7,02
10,06
25 °C
1,68
4,01
7,00
10,01
30 °C
1,68
4,01
6,99
9,97
35 °C
1,69
4,02
6,98
9,93
40 °C
1,69
4,03
6,97
9,89
45 °C
1,70
4,05
6,97
9,86
50 °C
1,71
4,06
6,96
9,83
FR-12 PH-mètre de laboratoire STARTER 5000
Mesure d'un échantillon
4.1.3 Mesure du pH
S'assurer que llectrode a é étalonnée avant la mesure. Après rinçage,
Placer l'électrode de pH dans l'échantillon.
Appuyer sur pour marrer la mesure du pH, lorsque le sultat final a é atteint, la
mesure en cours est terminée.
Remarque : Si le signal est constamment instable, il est recommandé d'utiliser le mode de
sultat final manuel afin d'éviter une mauvaise rétitivité.
4.1.4 Mesure en mV et Rel.mV
Appuyer sur pour passer sur ou sur mode ; appliquer la me
procédure que pour la mesure du pH afin de réaliser une mesure en mV ou en Rel.mV.
Si l'appareil est connecté à une électrode ORP, le sultat en mV correspond à la valeur du
potentiel d'oxydo-réduction, aussi connue sous le nom de valeur Redox.
4.2 Mesure de température
Pour une précision supérieure, il est recommandé d'utiliser une électrode de température intégrée
ou séparée.
Si une électrode de température est utilisée, ATC et la température de l'échantillon sont
affichés.
Si l'appareil ne détecte pas d'électrode de température, il passe automatiquement au mode
manuel de compensation de température et MTC apparaît. La température MTC doit être
finie.
PH-mètre de laboratoire STARTER 5000 FR-13
4.3 Impression
Le STARTER 5000 peut imprimer la mesure ou les données d'étalonnage avec le port RS232.
Après avoir correctement connecté l'imprimante (par exemple, OHAUS SF-40A) au pH-mètre,
procéder aux paramétrages suivants (voir 5.1) :
commander le ble de connexion du ST5000 (réf. 30059316) pour connecter les SF40A et
ST5000.
Paramétrage système ► Sortie, sélection Imprimante
Paramétrage système bit en bauds, sélectionner 9600.
Le format de l'impression des mesures est le suivant :
Date Heure Utilisateur ID échantillon
6-13-2015 14:20:22 Utilisateur01 Échantillon02
CapteurID Mode de résultat final MTC/ATC
Capteur02 Manuel ATC
Température Résultats
25,3 ºC 6,999 pH
Le format de l'impression des données d'étalonnage est le suivant :
Données d'étalonnage
CapteurID Numéro de série
Capteur02 14060012
Étalonnage Date Étalonnage Heure ATC/MTC
01-01-2015 14:20:33 ATC
Tampons mVcalage Pente
4,00 pH 176,94 mV
7,00 pH 0,05 mV 0,05 mV 98,22 %
Lors de la réalisation d'une mesure en continu, le format des données d'étalonnage imprimées est
le suivant :
Synchronisation de l'enregistrement des mesures
CapteurID Numéro de série
Capteur03 14060012
Date Heure Température Mesure
01-01-2015 14:20:25 23,5 ºC 6,998 pH
Date Heure Température Mesure
01-01-2015 14:20:30 23,5 ºC 6,999 pH
Date Heure Température Mesure
01-01-2015 14:20:35 23,5 ºC 7,000 pH
PH-mètre de laboratoire STARTER 5000 FR-14
5 CONFIGURATION
5.1 Paramétrage système
Le paramétrage système est utilisé pour finir les paramètres de l'appareil, et les informations
concernant ce dernier s'y trouvent aussi. Consulter la Section 3.4 pour des rérences.
5.1.1 Paramétrage système ► Mode Enregistrement
Le mode Enregistrement possède 2 options, l'une concerne l'enregistrement automatique de la
mesure du résultat final et l'autre concerne l'enregistrement manuel de la mesure du sultat final.
« Enregistrement automatique de la lecture du sultat final » signifie que si la mesure atteint le
sultat final (automatique, manuel ou temporisé), l'appareil enregistrera automatiquement la
lecture du résultat final.
L'icône d'enregistrement passera de noir à gris clair . La mesure d'un sultat
final ne peut être enregistrée qu'une seule fois.
5.1.2 Paramétrage système ► Sortie
Après avoir lectionné l'arrêt de la connexion, l'interface principale affichera l'icône
d'enregistrement ou affichera l'ine de sortie .
5.1.3 Paramétrage système ► Restauration des paramètres par défaut
La restauration selon la configuration usine comprend ce qui suit :
1. Mode GLP d'arrêt
2. Mode d'auto enregistrement avec arrêt automatique
3. Arrêt des mesures en continu
4. Arrêt de connexion (sortie)
5. Arrêt du rappel de l'étalonnage
6. La résolution est : 0,001 pH, 0,01mV
7. Mode de mesure : pH
8. Le groupe d'auto-tampons sera supprimé
9. bit en bauds : 9600
10. Luminosité de l'écran : 70 %
11. Arrêt de la protection de l’écran
12. MTC 25 ºC
13. sultat : Auto
14. Indicateur temporel du résultat final : 10 secondes
15. Supprimer toutes les données d'étalonnage
5.1.4 Paramétrage système ► Mot de passe de service
Non utilisé.
PH-mètre de laboratoire STARTER 5000 FR-15
5.2 Paramétrage des mesures
5.2.1 Paramétrage des mesures ► Mode de résultat final
sultat final automatique
Dans ce mode, l'appareil atteindra le sultat final et verrouillera automatiquement la mesure
conformément aux critères suivants de stabilité :
solution 0,1 pH : le signal de l'entrée du capteur peut ne pas être modifié de plus de 0,3 mV en
6 secondes.
solution 0,01 pH : le signal de l'entrée du capteur peut ne pas être modifié de plus de 0,1 mV
en 6 secondes.
solution 0,001 pH : le signal de l'entrée du capteur peut ne pas être modifié de plus de 0,03 mV
en 6 secondes.
sultat final temporel
Dans ce mode, l'utilisateur doit introduire la durée du sultat final, par exemple, 15 secondes.
Ceci signifie que 15 secondes après avoir débuté la mesure, l'appareil atteindra le résultat final.
sultat final Manuel
Dans ce mode, l'utilisateur doit décider si la mesure du signal est stable, pour appuyer ensuite sur
l'icône d'arrêt et atteindre unsultat final manuel.
5.2.2 Paramétrage des mesures ► Mode GLP de mesure
Le mode GLP impose à chaque mesure de recevoir l'ID utilisateur et l'ID échantillon. En l'absence
de l'un ou de l'autre, l'appareil affichera l'interface permettant à l'utilisateur d'introduire les
données requises.
5.2.3 Paramétrage des mesures ► Mesure en continu
Pendant une mesure en continu, l'utilisateur doit introduire l'intervalle temporel afin d'obtenir le
sultat de chaque mesure. Le Graphique de la mesure en continu se trouve ci-dessous.
L'axe des Y correspond à la valeur du pH. L'axe des X correspond au nombre de sultats de
mesures.
PH-mètre de laboratoire STARTER 5000 FR-16
5.3 Paramétrage de l'étalonnage
5.3.1 Paramétrage de l'étalonnage ►Groupe de tampons
Avant l'étalonnage, s'assurer que le groupe actuel de tampons est correct pour les solutions de
tampons. L'utilisateur peut sélectionner parmi 8 groupes de tampons pfinis.
L'utilisateur peut aussi établir des groupes de tampons auto-définis. La différence entre 2 points
d'étalonnage adjacent doit être supérieure à 0,99 pH.
5.3.2 Paramétrage de l'étalonnage ► Rappel de l'étalonnage
Entrer ce glage afin de finir la durée, depuis le dernier étalonnage, après combien d'heures
l'appareil doit rappeler la conduite d'un nouvel étalonnage. L'appareil affichera l'ine de rappel de
l'étalonnage et clignotera afin de rappeler à l'utilisateur d'effectuer un étalonnage.
5.4 Journal des données
Appuyer sur Données pour entrer sur l'interface du journal des données, notamment celles des
mesures et des étalonnages.
Vous pouvez supprimer ici toutes les données de mesures. La restauration des paramètres par
faut reste le seul moyen de supprimer les données d'étalonnage.
Les données peuvent être dirigées vers une clé USB (4 Go).
PH-mètre de laboratoire STARTER 5000 FR-17
6 MAINTENANCE
6.1 Message d’erreur
Lorsque la valeur introduite/le sultat de la mesure sont hors limites, ou si les données
d'étalonnage ne sont pas suffisamment bonnes, l'appareil affichera des informations d'erreur.
Si l'erreur ne peut pas être solue en utilisant le message d'erreur affiché, contacter le
représentant commercial d'Ohaus pour obtenir de l'aide.
6.2 Maintenance de l'appareil de mesure
Les instruments de la rie STARTER 5000 ne cessitent aucune maintenance autre qu'un
essuyage occasionnel avec un chiffon humide.
Attention : Suivre ces précautions de sécurité lors du nettoyage de l'instrument.brancher
l'instrument de sa prise d'alimentation.
Ne pas ouvrir l'instrument.
Ne laisser aucun liquide pétrer dans l'instrument.
Attention : L'enceinte est fabriquée en polystyrène-butadiène-acrylonitrile (ABS). Ce matériau est
attaqué par certains solvants organiques comme le toluène, le xylène et l'éthyle méthyle tone
(MEK). Tout déversement doit être immédiatement asséché.
6.3 Maintenance de l'électrode
S'assurer que l'électrode est remplie d'une solution électrolytique. Toujours stocker l'électrode
conformément au manuel d'instructions de l'électrode ; ne pas la laisser s'assécher.
Si la ponse de l'électrode devient lente ou si la pente n'est pas suffisamment bonne, essayer ce
qui suit :
o Tremper l'électrode dans de HCl 0,1M pendant plus de 8 heures.
o Pour des contaminants huileux ou gras, graisser la membrane avec de la laine de
coton trempée dans de l'acétone ou dans une solution savonneuse.
Après avoir traité l'électrode, un nouvel étalonnage doit être réalisé. Si la pente de llectrode n'est
toujours pas bonne, l'électrode doit être remplacée.
6.4 Informations
Appuyer sur l'icône Informations pour entrer sur l'interface d'informations. L'utilisateur peut ici
introduire des informations pertinentes en appuyant sur la zone avec « » en bas et à gauche,
telles que la date d'achat du capteur, etc.
FR-18 PH-mètre de laboratoire STARTER 5000
7 DONNÉES TECHNIQUES
7.1 Spécifications
Les données techniques sont valides dans les conditions ambiantes suivantes :
Utilisation en intérieur seulement
Altitude : Jusqu'à 2000 m
Plage de température spécifiée : 5 à 40°C
Humidité : humidité relative maximum à 80 % avec des températures jusqu'à 30°C diminuant
linéairement à 50 % d'humidité relative à 40°C
Variations secteur jusqu'à ± 10 % de la tension nominale
Installation de catégorie II
Degré de pollution : 2
Fonctionnement assuré à température ambiante entre 5 et 40°C
Modèle
STARTER 5000
Plage de mesure
-2,000…+ 20,000 pH
–2000.00…+2000.00 mV
-30 °C…+130 °C
solution
0,1/0,01/0,001 pH
1/0,1/0,01 mV
0,1 °C
Limites d'erreur
± 0,002 pH
± 0,03 % FS mV
± 0,1 °C
Étalonnage
1-9 points
8 groupes de tampons préfinis + 1 groupe de tampons
auto-définis
moire
1 000 mesures
Données d'étalonnage 10*10
Alimentation
Adaptateur CA, entrée 100-240 V~0,3 A input,
sortie 9 V 1,12A
Dimensions/poids
Environ 210 x 170 x 80 mm/0,65 kg
Poids/dimensions d'expédition
Environ 372 x 310 x 185 mm/2,3 kg
Affichage
ACL tactile avec rétroéclairage
Saisie
BNC, impédance > > 3 x 10e+12 Ω
Cinch, NTC 30 kΩ
Compensation de température
ATC et MTC
Enceinte
ABS
PH-mètre de laboratoire STARTER 5000 FR-19
8 CONFORMITÉ
Ce produit est conforme à la directive EMC 2004/108/EC et à la directive de
basse tension 2006/95/EC. Sa Déclaration de conformité est disponible en ligne
sur europe.ohaus.com/europe/en/home/support/compliance.aspx.
Conformément à la directive européenne 2002/96/CE sur les déchets
d’équipements électriques et électroniques (WEEE), cet appareil ne doit pas
être mis au rebut comme des ordures ménagères. Ceci est également valable
pour les pays en dehors de l’UE, selon leurs spécifications particulières. La
directive 2006/66/EC sur les batteries a introduit de nouveaux impératifs depuis
septembre 2008 sur le retrait des batteries d'un équipement destiné au rebut
dans tous les États membres de l'Union européenne. Pour être conforme à
cette directive, cet appareil a été conçu pour un retrait sans danger des
batteries aussit sa fin de vie par une installation de traitement des déchets.
Veuillez mettre au rebut ce produit conforment à la réglementation locale au
point de collecte spécifié pour les équipements électriques et électroniques. En
cas de questions, veuillez contacter l’autorité responsable ou le distributeur
auprès duquel cet appareil a é acheté. Si cet appareil change de proprtaire
(pour des raisons personnelles ou professionnelles), cette consigne doit lui être
communiquée. Les instructions de mise au rebut en Europe sont disponibles en
ligne sur europe.ohaus.com/europe/en/home/support/weee.aspx. Nous vous
remercions de votre contribution à la protection de l’environnement.
Remarque FCC : Cet équipement a é testé et déclaré conforme aux limites des appareils
numériques de classe A, selon la Section 15 des règles du FCC (Commission férale des
communications). Ces limites ont pour objectif de fournir une protection raisonnable contre
des interférences dangereuses lorsque l'équipement est utilisé dans un environnement
commercial. Cet équipement gére, utilise et peut rayonner des hautes fréquences et s'il
n'est pas installé et utilisé conforment au guide de l'utilisateur, il peut gérer des
interférences préjudiciables aux communications radio. L'utilisation de cet équipement dans
une zone résidentielle risque de gérer des interférences préjudiciables, auquel cas
l'utilisateur se verra dans l'obligation de rectifier la situation à ses frais.
Remarque d’Industrie Canada
Cet appareil nurique de classe A est conforme à la norme canadienne ICES-001.
Enregistrement ISO 9001
En 1994, le Bureau Veritus Quality International (BVQI) a octroyé la certification
d'enregistrement ISO 9001 à Ohaus Corporation, États-Unis d'Amérique, confirmant que le
système de gestion de la qualité d'OHAUS était conforme aux spécifications normalisées de
l'ISO 9001. Le 21 juin 2012, Ohaus Corporation, États-Unis d'Amérique, a été-enregistrée
à la norme ISO 9001:2008.
FR-20 PH-mètre de laboratoire STARTER 5000
9 GROUPES DE TAMPONS
Le STARTER 5000 corrige automatiquement la dépendance à la température des valeurs de pH du groupe de
tampons qui sont données dans les tableaux suivants. Le groupe de tampons doit être choisi avant
l'étalonnage.
JJG119 ((Réf. 25°C, 0,01pH)
Temp °C 1,68 4,00 6,86 9,18 12,46
5 1,67 4,00 6,95 9,39 13,21
10 1,67 4,00 6,92 9,33 13,01
15 1,67 4,00 6,90 9,28 12,82
20 1,68 4,00 6,88 9,23 12,64
25 1,68 4,00 6,86 9,18 12,46
30 1,68 4,01 6,85 9,14 12,29
35 1,69 4,02 6,84 9,11 12,13
40 1,69 4,03 6,84 9,07 11,98
45 1,70 4,04 6,83 9,04 11,83
50 1,71 4,06 6,83 9,02 11,70
Normes des États-Unis (Réf, 25 °C)
Temp °C 1,68 4,01 7,00 10,01
5 1,67 4,01 7,09 10,25
10 1,67 4,00 7,06 10,18
15 1,67 4,00 7,04 10,12
20 1,68 4,00 7,02 10,06
25 1,68 4,01 7,00 10,01
30 1,68 4,01 6,99 9,97
35 1,69 4,02 6,98 9,93
40 1,69 4,03 6,97 9,89
45 1,70 4,05 6,97 9,86
50 1,71 4,06 6,96 9,83
Norme européenne (Réf, 25 °C)
Temp °C 2,00 4,01 7,00 9,21 11,00
5 2,02 4,01 7,09 9,45 11,72
10 2,01 4,00 7,06 9,38 11,54
15 2,00 4,00 7,04 9,32 11,36
20 2,00 4,00 7,02 9,26 11,18
25 2,00 4,01 7,00 9,21 11,00
30 1,99 4,01 6,99 9,16 10,82
35 1,99 4,02 6,98 9,11 10,64
40 1,98 4,03 6,97 9,06 10,46
45 1,98 4,04 6,97 9,03 10,28
50 1,98 4,06 6,97 8,99 10,10
PH-mètre de laboratoire STARTER 5000 FR-21
MERCK (Réf, 20 °C)
Temp °C 2,00 4,00 7,00 9,00 12,00
5 2,01 4,04 7,07 9,16 12,41
10 2,01 4,02 7,05 9,11 12,26
15 2,00 4,01 7,02 9,05 12,10
20 2,00 4,00 7,00 9,00 12,00
25 2,00 4,01 6,98 8,95 11,88
30 2,00 4,01 6,98 8,91 11,72
35 2,00 4,01 6,96 8,88 11,67
40 2,00 4,01 6,95 8,85 11,54
45 2,00 4,01 6,95 8,82 11,44
50 2,00 4,00 6,95 8,79 11,33
DIN (19267) (Réf, 25 °C)
Temp °C 1,09 4,65 6,79 9,23 12,75
5 1,08 4,67 6,87 9,43 13,63
10 1,09 4,66 6,84 9,37 13,37
15 1,09 4,66 6,82 9,32 13,16
20 1,09 4,65 6,80 9,27 12,96
25 1,09 4,65 6,79 9,23 12,75
30 1,10 4,65 6,78 9,18 12,61
35 1,10 4,65 6,77 9,13 12,45
40 1,10 4,66 6,76 9,09 12,29
45 1,10 4,67 6,76 9,04 12,09
50 1,11 4,68 6,76 9,00 11,98
JIS Z 8802 (Réf, 25 °C)
Temp °C 1,679 4,008 6,865 9,180
5 1,668 3,999 6,951 9,395
10 1,670 3,998 6,923 9,332
15 1,672 3,999 6,900 9,276
20 1,675 4,002 6,881 9,225
25 1,679 4,008 6,865 9,180
30 1,683 4,015 6,853 9,139
35 1,688 4,024 6,844 9,102
40 1,694 4,035 6,838 9,068
45 1,700 4,047 6,834 9,038
50 1,707 4,060 6,833 9,011
DIN(19266) / NIST (Réf, 25 °C)
Temp °C 1,680 4,008 6,865 9,183 12,454
5 1,668 4,004 6,950 9,392 13,207
10 1,670 4,001 6,922 9,331 13,003
15 1,672 4,001 6,900 9,277 12,810
20 1,676 4,003 6,880 9,228 12,627
25 1,680 4,008 6,865 9,183 12,454
30 1,685 4,015 6,853 9,144 12,289
35 1,691 4,026 6,845 9,110 12,133
40 1,697 4,036 6,837 9,076 11,984
45 1,704 4,049 6,834 9,046 11,841
50 1,712 4,064 6,833 9,018 11,705
FR-22 PH-mètre de laboratoire STARTER 5000
JJG119 (Réf, 25 °C, 0,001pH)
Temp °C 1680 4,003 6,864 9,182 12460
5 1,669 3,999 6,949 9,391 13,210
10 1,671 3,996 6,921 9,330 13,011
15 1,673 3,996 6,898 9,276 12,820
20 1,676 3,998 6,879 9,226 12,637
25 1,680 4,003 6,864 9,182 12,460
30 1,684 4,010 6,852 9,142 12,292
35 1,688 4,019 6,844 9,105 12,130
40 1,694 4,029 6,838 9,072 11,975
45 1,700 4,042 6,834 9,042 11,828
50 1,706 4,055 6,833 9,015 11,697
PH-mètre de laboratoire STARTER 5000 FR-23
GARANTIE LIMITÉE
Ohaus garantit que ses produits sont exempts de défauts matériels et de fabrication à compter de
la date de livraison pendant toute la durée de la garantie, Pendant la période de garantie, Ohaus
parera ou, à sa convenance, remplacera sans frais toutes les pièces déterminées défectueuses,
sous réserve que le produit soit retourné, fret payé d'avance, à Ohaus,
Cette garantie ne s'applique pas si le produit a subi des dommages suite à un accident ou un
susage, a é exposé à des matériaux radioactifs ou corrosifs, contient des matériaux
étrangers ayant pétré à l'intérieur ou après une maintenance ou des modifications apportées
par quiconque autre qu’Ohaus, En l’absence d’une carte d'enregistrement de garantie dûment
remplie, la période de garantie commence à la date de l'exdition au revendeur agréé, Aucune
autre garantie expresse ou implicite n'est offerte par Ohaus Corporation, En aucun cas, Ohaus
Corporation ne peut être tenu responsable des dommages indirects,
Dans la mesure les lois régissant les garanties varient d'un État à un autre et d'un pays à un
autre, veuillez contacter Ohaus ou votre représentant local agréé Ohaus pour de plus amples
informations,
EN-22 STARTER 5000 Bench pH Meter
Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054 USA
Tel: (973) 377-9000
Fax: (973) 944-7177
With offices worldwide/ Con oficinas en todo el mundo/ Avec des bureaux
dans le monde entier
www.ohaus.com
P/N 30129898 C© 2014 Ohaus Corporation, all rights reserved/ todos los
derechos reservados/ tous droits réservés.
*30129898*

Transcripción de documentos

STARTER 5000 Bench pH Meter Instruction Manual STARTER 5000 Medidor de pH de banco Manual de instrucciones STARTER 5000 PH-mètre de laboratoire Manuel d’instructions TABLE OF CONTENTS INTRODUCTION ................................................................1 1 1.1 1.2 2 Definition of Signal Warnings and Symbols ................................. 1 Safety Precautions ........................................................................... 2 INSTALLATION .................................................................3 2.1 2.2 2.3 2.4 3 Package contents............................................................................. 3 Installing the stand-alone electrode holder ................................... 4 Installing the power adapter ........................................................... 4 Connect the pH electrodes ............................................................. 5 DISPLAY AND ICONS .......................................................5 3.1 3.2 3.3 3.4 4 Turn on the meter ............................................................................. 5 Turn off the meter............................................................................. 6 Icons .................................................................................................. 6 Menu directory ................................................................................. 9 CALIBRATION, MEASUREMENT AND PRINT .............. 10 4.1 4.1.1 4.1.2 4.2 4.2.1 4.2.2 4.3 4.4 5 Calibration ...................................................................................... 10 Buffer group ..................................................................................... 10 Performing calibration .......................................................................11 Sample measurement .................................................................... 12 pH measurement ............................................................................. 12 mV and Rel.mV measurement ......................................................... 12 Temperature measurement ........................................................... 12 Printing............................................................................................ 13 SETUP ............................................................................. 14 5.1 5.1.1 5.1.2 5.1.3 5.1.4 5.2 5.2.1 5.2.2 5.2.3 5.3 5.3.1 5.3.2 5.4 6 System Setting ............................................................................... 14 System setting ►Save Mode .......................................................... 14 System Setting ►Output ................................................................. 14 System Setting ► Restore Factory Setup ....................................... 14 System Setting ► Service Password .............................................. 14 Measurement Setting ..................................................................... 15 Measurement Setting ►Endpoint Mode .......................................... 15 Measurement Setting ►GLP Measurement Mode .......................... 15 Measurement Setting ►Continuous Measure ................................. 15 Calibration Setting ......................................................................... 16 Calibration Setting ►Buffer Group .................................................. 16 Calibration Setting ►Calibration Reminder ..................................... 16 Data Log .......................................................................................... 16 MAINTENANCE ............................................................... 17 6.1 6.2 Error message ................................................................................ 17 Meter maintenance ........................................................................ 17 6.3 6.4 7 TECHNICAL DATA .......................................................... 18 7.1 8 9 Electrode maintenance .................................................................. 17 Information ..................................................................................... 17 Specifications ................................................................................. 18 COMPLIANCE ................................................................. 19 BUFFER GROUPS .......................................................... 20 STARTER 5000 Bench pH Meter 1 EN-1 INTRODUCTION Thank you for choosing this OHAUS product. Please read the manual completely before using the STARTER 5000 bench pH meter to avoid incorrect operation. STARTER 5000 is a 0.001pH bench meter with a color touch-screen LCD. It’s a new generation of intuitive and easy-to-use bench meters for reliable measurements. A few of the many existing features are:  Bi-lingual graphical user interface  Auto, manual time endpoint which can assist users to freeze the stable reading value to meet different customer requirements  8 buffer groups (US, EU, NIST and JJG) with 1 self-defined buffer  IP54 water and dust protection 1.1 Definition of Signal Warnings and Symbols Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results. Signal Words WARNING CAUTION Attention Note For a hazardous situation with medium risk, possibly resulting in injuries or death if not avoided. For a hazardous situation with low risk, resulting in damage to the device or the property or in loss of data, or injuries if not avoided. For important information about the product For useful information about the product Warning Symbols General Hazard Electrical Shock Hazard Explosion Hazard Corrosive Hazard Alternating Current Direct Current EN-2 1.2 STARTER 5000 Bench pH Meter Safety Precautions CAUTION Read all safety warnings before installing, making connections, or servicing this equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain all instructions for future reference.           Verify that the AC adapter's input voltage range and the plug type are compatible with the local AC power supply. Make sure that the power cord does not pose a potential obstacle or tripping hazard. Use the equipment only in dry locations. Dry off any liquid spills immediately. The instrument is not watertight. Use only approved accessories and peripherals. Operate the equipment only under ambient conditions specified in these instructions. Disconnect the equipment from the power supply when cleaning. Do not operate the equipment in hazardous or unstable environments. Service should only be performed by authorized personnel. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. WARNING Do not use this instrument in an environment subject to explosion hazards. The housing of the instrument is not gas tight (explosion hazard due to spark formation, corrosion caused by the ingress of gases). CAUTION When using chemicals and solvents, comply with the instructions of the producer and the general lab safety rules. STARTER 5000 Bench pH Meter 2 EN-3 INSTALLATION Carefully unpack the box. 2.1 Package contents The model ST5000-B (basic package) has the following items: ST5000-B ST5000 meter ST5000 In use cover Stand-alone electrode holder 4GB USB flash drive Buffer powder set (4.01; 7.00; 10.00) Power supply Units 1 1 1 set 1 1 set 1 set In addition to ST5000-B content, the ST5000-F package also includes the following: ST350 3-in-1 glass refillable pH electrode 1 Each pH buffer powder should be dissolved in 250ml pure water or deionized water in a volumetric flask. You also can order the bottled buffer solution from Ohaus. More electrodes, solutions, accessories and spare parts, please contact Ohaus for more details. Model ST350 ST310 ST210 ST320 ST260 ST230 STSURF STMICRO5 STMICRO8 STTEMP30 STORP1 STORP2 Description 3-in-1 glass refillable pH Electrode 3-in-1 plastic refillable pH Electrode 2-in-1 plastic refillable pH Electrode 3-in-1 plastic gel pH Electrode(no need to fill) 2-in 1 glass double salt bridge pH Electrode 2-in-1 glass muddy sample pH Electrode 2-in-1 flat surface pH Electrode 2-in-1 glass micro sample pH Electrode 2-in-1 glass micro sample pH Electrode Temperature Electrode Gel plastic ORP electrode Refillable glass ORP electrode Buffer powder set (4.01; 7.00; 10.0) Buffer pH1.68 250ml Buffer pH4.01 250ml Buffer pH7.00 250ml Buffer pH10.01 250ml Buffer pH12.45 250ml pH electrode Reference Electrolyte pH electrode protection solution Electrode holder stand alone ST5000 In use cover SP Cable Adapter RS232 F-M SF40A (for ST5000) P/N 30129354 83033965 83033966 83033967 30129357 83033968 30129470 30087566 30087569 83033970 30038555 30038553 83033971 30100424 30100425 30100427 30100429 30100440 30059255 30059256 30058733 30129897 30059316 EN-4 2.2 STARTER 5000 Bench pH Meter Installing the stand-alone electrode holder Install the electrode arm on the base: After adjusting the tension knob to some extent, the upper arm can be move up and down. When the pH electrode is placed on the arm and calibration or measurement is being performed, the storage bottle of pH electrode can be placed into the base hole for easy storage: 2.3 Installing the power adapter Insert the adapter plug into the power adapter slot. CAUTION Only use an adapter plug that is compatible with the mains power receptacle. Make sure that the adapter plug is fully seated in the power adapter slot before plugging it into the mains power receptacle. STARTER 5000 Bench pH Meter 2.4 EN-5 Connect the pH electrodes There are 2 sockets for the electrode. “pH” socket (BNC); “TEMP.” Socket (Cinch). For ST350 3-in-1 electrode, connect both the BNC and Cinch socket. For 2-in-1 pH electrodes (e.g. ST230), only connect the BNC. 3 DISPLAY AND ICONS 3.1 Turn on the meter STARTER 5000 is a color touch screen 0.001pH bench meter. It does not have any buttons. After connecting with power, the meter will turn on, if meter in screen protection, you can tap the screen to turn on the screen. Note: If customer did not operate the meter for more than 6 hours, the meter will completely shut off automatically. You cannot turn on the meter by tapping the screen any more. You need to reconnect power to turn on the meter. Select the language and then press Main Interface to enter the main interface. EN-6 3.2 STARTER 5000 Bench pH Meter Turn off the meter To turn off the meter, please tap the turn-off button the turn-off interface. Then tap main interface again. in the top-left of main interface to enter to turn-off the meter or tap to go back to Turn-off interface 3.3 Icons User ID Sample ID Information icon Date Time pH measurement mode. Shift between mV and Rel.mV mode or Signal icon, when it is blinking that means the meter is receiving signal (in measurement or calibration). Endpoint icons, auto-endpoint, manual-endpoint and time-endpoint. When in manual-endpoint, if the signal is stable, the meter will show the stable signal icon. Tap the stop icon Signal stable icon Temperature icon. to reach manual endpoint. STARTER 5000 Bench pH Meter EN-7 Information, measure and calibrate tap area The information area is also the shortcut to enter the setting interface: "Endpoint:" Tap this area to enter the interface to set the endpoint mode. Available modes are Auto, Manual and Time endpoint. The current endpoint mode is displayed. "Sensor ID:" Tap this area to enter the interface to select, set or change the sensor ID. The current sensor ID is displayed. "Sensor Cond:" Tap this area to enter the sensor calibration data interface. The current sensor performance icon (judged by calibration data) is displayed. There are 3 sensor condition icons: Slope: more than 95% Slope: 90-95% and offset: ± (0-15) mV Electrode condition is good or offset: ± (15-35) mV Electrode condition is acceptable Slope: less than 90% or offset: ± (35-60) mV Electrode condition is not good or needs cleaning Calibration reminder icon. Set the time you want the meter to alert you to do calibration from last calibration. When the time is reached, this icon will be displayed and blink to remind that it is time to perform a calibration. “Buffers:” This area displays the current buffer group, tap this area to enter the interface to select the buffer group. Start icon. Tap this icon and the meter will begin measurement, the signal icon will blink. The start icon area will show the stop icon Calibration icon. Tap this icon to start calibration. Save icon. After saving the current measurement result (endpoint reached), this icon color will change from black to light gray , indicating that the reading has already been saved and cannot be saved twice. Output icon. Output data to external device such as a printer. If “Shut off connection” is chosen in System Setting ►Output, the save icon will be displayed .Otherwise the output icon will be displayed in the same position. EN-8 STARTER 5000 Bench pH Meter Return to main interface Return to previous menu Enter, confirm or apply current setup, finish setting Edit Page up Page down Delete Increase (light) Decrease (light) Graph display (for the measurement or calibration data) Save Output to external device (USB flash drive, printer) Full keyboard Next calibration point Current Calibration point again End of calibration STARTER 5000 Bench pH Meter 3.4 Menu directory The menus of first and secondary directory for STARTER 5000: Date Time Save Mode Output System Setting Screen Brightness Baud Rate Screen Protection Restore Factory Setup Instrument Information Service Password Resolution MTC Temp. Input pH/mV/Rel.mV Mode Measurement Setting Continuous Measure Endpoint Mode ID setting GLP Measurement Mode Calibration Setting Data Buffer Group Calibration Reminder Measurement Data Calibration Data EN-9 EN-10 4 STARTER 5000 Bench pH Meter CALIBRATION, MEASUREMENT AND PRINT The meter display will turn off automatically according to the time set in the menu System Setting ►Screen Protection ►LCD Auto-off. To turn on the screen again, tap the screen.(See 3.1) A complete pH measurement process should follow the procedure: a) Rinse the pH electrode and connect to meter b) Prepare the buffer and calibrate the pH electrode c) Prepare the sample d) Conduct pH measurement of the sample e) Record measurement data f) Rinse the pH electrode and store Prepare the pH electrode: the pH electrode should be rinsed with pure water before and after usage. Check if the electrode is physically damaged. (Be careful with the glass bulb) The pH electrode should be stored in the storage bottle. After the pH electrode has been put into the sample or buffer solution, stir several seconds then wait 30 to 60 seconds for the signal to be stable, and then press the button to operate (Calibration or measurement). WARNING Do not operate the equipment in hazardous environments. The equipment is not explosion protected. CAUTION When using chemicals and solvents, comply with the instructions of the chemical producer and the general lab safety rules. 4.1 Calibration 4.1.1 Buffer group STARTER 5000 can perform 1-5 point calibrations for predefined buffer groups. For self-defined buffer group, it can do 1-9 point calibration. Before calibration, make sure the current buffer group is the group wanted. When using a new pH electrode, a calibration should be performed before any measurement. If a pH electrode is not used for a few days, a calibration should be performed before next usage. Calibration is to display the right pH value when the meter receive the mV value signal from the pH electrode. Slope: the linear coefficient between mV and pH according to theoretical value (e.g. 59.16mV/pH @ 25°C means 100% slope); Offset: the mV value when the pH value is 7.00. (Theoretical value is 0 mV) STARTER 5000 Bench pH Meter EN-11 STARTER 5000 will give an electrode condition icon to show the performance of the pH electrode according to the calibration data. Normally, if we get sad face icon after a proper calibration, you may need to replace the pH electrode with a new one. STARTER 5000 automatically corrects for the temperature dependence of the buffer pH values. E.g. for US standard buffer group "1.68, 4.01, 7.00, 10.01" reference the following table: 5°C 1.67 4.01 7.09 10.25 10°C 1.67 4.00 7.06 10.18 15°C 1.67 4.00 7.04 10.12 20°C 1.68 4.00 7.02 10.06 25°C 1.68 4.01 7.00 10.01 30°C 1.68 4.01 6.99 9.97 35°C 1.69 4.02 6.98 9.93 40°C 1.69 4.03 6.97 9.89 45°C 1.70 4.05 6.97 9.86 50°C 1.71 4.06 6.96 9.83 For example, with calibration buffer 10.01 at 15°C, the calibration result should be 10.12 pH. 4.1.2 Performing calibration After rinsing the pH electrode, put the electrode into the first buffer, stir for 10s then press to start the calibration. When in calibration, the signal icon blinks. When the endpoint has been reached, appears while are now 3 options: disappears and current point calibration is finished, There Quit the calibration and go back to main interface Finish current point calibration, go to calibration result interface Do the next point calibration Note: During calibrating process, the ST5000 is always in Auto Endpoint Mode. Note: The use of 3-in-1 pH electrode which include the temperature sensor or a separate temperature electrode is recommended. When using the MTC mode, enter the correct temperature value and keep all buffer and sample solutions at the set temperature. EN-12 STARTER 5000 Bench pH Meter 4.2 Sample measurement 4.2.1 pH measurement Be sure the electrode has been calibrated before measurement. After rinsing,  Place the pH electrode in the sample.  Press to start the pH measurement, when the endpoint has been reached, the current measurement is finished. Note: If the signal is not stable all the time, it is recommended to use manual endpoint mode to avoid bad repeatability. 4.2.2 mV and Rel.mV measurement Press to shift to or mode; follow the same procedure as for pH measurement to perform mV or Rel.mV measurement. If the meter is connected to an ORP electrode, the mV result is the value of Oxidation-Reduction Potential value; also called Redox value. 4.3 Temperature measurement For better accuracy, we recommend to use either a built-in or a separate temperature electrode.  If a temperature electrode is used, ATC and the sample temperature are displayed.  If the meter does not detect a temperature electrode, it automatically switches to the manual temperature compensation mode and MTC appears. MTC temperature should be set. Note: STARTER 5000 accepts NTC 30 kΩ temperature sensor. For ATC (Auto Temperature Compensation) or MTC (Manual Temperature Compensation), the Temperature Compensation only corrects for the change in the output of the electrode, not for the change in the actual solution. That means the meter corrects the signal (mV) from the pH electrodes to get a more accurate pH value according to the real temperature. STARTER 5000 Bench pH Meter 4.4 EN-13 Printing STARTER 5000 can print the measurement or calibration data through RS232. After the printer (e.g. OHAUS SF-40A) is connected to the meter properly, make the following settings(See 5.1): please order the connect cable for ST5000 (P/N 30059316) to connect SF40A and ST5000. System Setting ► Output , select Printer System Setting ► Baud Rate, select 9600. The format for the print-out measurement is: Date Time User SampleID 6-13-2015 14:20:22 User01 Sample02 SensorID Endpoint mode MTC/ATC Sensor02 Manual ATC Temperature Result 25.3°C 6.999pH The format for the print-out calibration data is: Calibration Data SensorID Serial Number Sensor02 14060012 Cal. Date Cal. Time ATC/MTC 01-01-2015 14:20:33 ATC Buffers mV Offset Slope 4.00pH 176.94mV 7.00pH 0.05mV 0.05mV 98.22% When performing continuous measurement, the format for the print-out calibration data is: Timing Measure Record SensorID Serial Number Sensor03 14060012 Date Time Temperature 01-01-2015 14:20:25 23.5°C Date Time Temperature 01-01-2015 14:20:30 23.5°C Date Time Temperature 01-01-2015 14:20:35 23.5°C Measurement 6.998pH Measurement 6.999pH Measurement 7.000pH EN-14 STARTER 5000 Bench pH Meter 5 SETUP 5.1 System Setting System setting is used to set meter parameters, the meter information could also be found in it. Please see section 3.4 for reference. 5.1.1 System setting ►Save Mode Save mode has 2 options, one is Auto Save Endpoint Reading, another is Manual Save Endpoint Reading. Auto Save Endpoint Reading means that when the measurement reaches the endpoint (auto, manual or time endpoint), the meter will save the endpoint reading automatically. The Save icon saved only once. 5.1.2 will change from black to light grey ; An endpoint reading can be System Setting ►Output When selecting shut off connection, the main interface will display the save icon display the Output Icon 5.1.3 . System Setting ► Restore Factory Setup Restoring the factory setup includes the following: 1. Shut off GLP mode 2. Shut off Auto Save mode 3. Shut off continuous measure 4. Shut off (output) connection 5. Shut off calibration Reminder 6. The resolution is: 0.001pH, 0.01mV 7. Measurement mode: pH 8. Self buffer group will be deleted 9. Baud rate: 9600 10. Brightness of screen: 70% 11. Shut off Screen Protection 12. MTC: 25°C 13. Endpoint: Auto 14. Time endpoint: 10 Second 15. Delete all the calibration data 5.1.4 Not used. System Setting ► Service Password , or STARTER 5000 Bench pH Meter 5.2 Measurement Setting 5.2.1 Measurement Setting ►Endpoint Mode EN-15 Auto Endpoint In this mode, the meter will reach the endpoint and lock the reading according to the following stability criteria automatically: 0.1pH resolution: the signal of the sensor input may not change by more than 0.3mV in 6 seconds. 0.01pH resolution: the signal of the sensor input may not change by more than 0.1mV in 6 seconds. 0.001pH resolution: the signal of the sensor input may not change by more than 0.03mV in 6 seconds. Time Endpoint In this mode, the user needs to input the endpoint time, e.g. 15 seconds. This means after starting the measurement for 15s, the meter will reach the endpoint. Manual Endpoint In this mode, the user needs to decide when the signal reading is stable, then press the stop icon to reach manual endpoint. 5.2.2 Measurement Setting ►GLP Measurement Mode GLP mode requires each measurement to have the User ID and Sample ID. If there is no sample ID or user ID, the meter will show the interface to let user input the required data. 5.2.3 Measurement Setting ►Continuous Measure During continuous measure, the user needs to input the time interval to get each reading result. Below is the Graph of continuous measure. The Y axis is pH value. The x axis is the Number of measurement result. EN-16 STARTER 5000 Bench pH Meter 5.3 Calibration Setting 5.3.1 Calibration Setting ►Buffer Group Before calibration, please make sure the current buffer group is correct for your buffer solutions. User can select from the 8 predefined buffer groups. User can also set self-defined buffer groups. The adjacent 2 calibration point difference should be larger than 0.99pH. 5.3.2 Calibration Setting ►Calibration Reminder Enter this setting to set the time, from the last calibration, after how many hours the meter should remind to perform calibration again. The meter will show the calibration reminder icon blink, to remind user to perform calibration. 5.4 and Data Log Tap Data to get into the Data log interface, including measurement data and calibration data. You can delete all the measurement data here. The only way to delete calibration data is to restore factory setup. The data can be output to a USB Flash Drive (≤4GB). STARTER 5000 Bench pH Meter 6 MAINTENANCE 6.1 Error message EN-17 When the input value/measurement result is out of range or calibration data is not good enough, the meter will display error information. If the error cannot be solved by the error message displayed, please contact Ohaus sales representative for further assistance. 6.2 Meter maintenance The STARTER 5000 series instruments do not require any maintenance other than occasional wiping with a damp cloth. Attention: Follow these safety precautions when cleaning the instrument  Disconnect the instrument from the power supply  Do not open the instrument  Do not allow liquids to enter the instrument Attention: The housing is made of acrylonitrile butadiene styrene (ABS). This material is attacked by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage should be immediately wiped off. 6.3 Electrode maintenance Make sure the electrode is filled with electrolyte solution. Always store the electrode according to the electrode instruction manual and do not allow it to dry out. If the electrode response becomes sluggish or the slope is not good enough, try the following: o Soak the electrode in 0.1M HCl for more than 8 hours. o For fat or oil contaminant, degrease the membrane with cotton wool soaked in either acetone or a soap solution. After electrode treatment, a new calibration should be performed. If the electrode slope is still not good, the electrode might need to be replaced. 6.4 Information Tap the Information icon to enter the information interface. Here the user can input related information by tapping the area with " " in the bottom left such as sensor purchase date, etc. EN-18 STARTER 5000 Bench pH Meter 7 TECHNICAL DATA 7.1 Specifications The technical data is valid under the following ambient conditions:  Indoor use only  Altitude: Up to 2000 m  Specified Temperature range: 5°C to 40°C  Humidity: maximum relative humidity 80 % for temperatures up to 30°C decreasing linearly to 50% relative humidity at 40°C  Mains supply voltage fluctuations: up to ±10% of the nominal voltage  Installation category II  Pollution degree: 2  Operability is assured at ambient temperatures between 5°C to 40°C Model STARTER 5000 Measuring range -2.000…20.000 pH –2000.00…2000.00 mV -30°C …130°C Resolution 0.1/0.01/0.001 pH 1/0.1/0.01 mV 0.1°C Error limits ± 0.002 pH ± 0.03% FS mV ± 0.1°C Calibration 1-9 points 8 predefined buffer group+1 self-defined buffer group Memory 1000 measurements 10*10 calibration data Power supply 100-240V~0.3A input, 9V Size/weight Approximately 210 x 170 x 80 mm / 0.65 kg Ship dimension/weight Approximately 372 x 310 x 185 mm / 2.30 kg Display Touch screen liquid crystal with backlight Input BNC, impedance > 3 x 10e+12 Ω Cinch, NTC 30 kΩ Temperature-compensation ATC & MTC Housing ABS 1.12A output AC Adapter STARTER 5000 Bench pH Meter 8 EN-19 COMPLIANCE This product conforms to the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC. The Declaration of Conformity is available online at europe.ohaus.com/europe/en/home/support/compliance.aspx. In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. The Batteries Directive 2006/66/EC introduces new requirements from September 2008 on removability of batteries from waste equipment in EU member States. To comply with this Directive, this device has been designed for safe removal of the batteries at end-of-life by a waste treatment facility. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. Disposal instructions in Europe are available online at europe.ohaus.com/europe/en/home/support/weee.aspx. Thank you for your contribution to environmental protection. FCC Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Industry Canada Note This Class A digital apparatus complies with Canadian ICES-001 ISO 9001 Registration In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the OHAUS quality management system is compliant with the ISO 9001 standards requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to the ISO 9001:2008 standard. EN-20 9 STARTER 5000 Bench pH Meter BUFFER GROUPS STARTER 5000 automatically correct for the temperature dependence of the buffer group pH value given in the following tables. Please choose the buffer group you want before calibration. JJG119 ((Ref. 25°C, 0.01pH) Temp °C 1.68 4.00 5 1.67 4.00 10 1.67 4.00 15 1.67 4.00 20 1.68 4.00 25 1.68 4.00 30 1.68 4.01 35 1.69 4.02 40 1.69 4.03 45 1.70 4.04 50 1.71 4.06 6.86 6.95 6.92 6.90 6.88 6.86 6.85 6.84 6.84 6.83 6.83 9.18 9.39 9.33 9.28 9.23 9.18 9.14 9.11 9.07 9.04 9.02 US standard (Ref. 25°C) Temp °C 1.68 4.01 5 1.67 4.01 10 1.67 4.00 15 1.67 4.00 20 1.68 4.00 25 1.68 4.01 30 1.68 4.01 35 1.69 4.02 40 1.69 4.03 45 1.70 4.05 50 1.71 4.06 7.00 7.09 7.06 7.04 7.02 7.00 6.99 6.98 6.97 6.97 6.96 10.01 10.25 10.18 10.12 10.06 10.01 9.97 9.93 9.89 9.86 9.83 Europe standard (Ref. 25°C) Temp°C 2.00 4.01 7.00 5 2.02 4.01 7.09 10 2.01 4.00 7.06 15 2.00 4.00 7.04 20 2.00 4.00 7.02 25 2.00 4.01 7.00 30 1.99 4.01 6.99 35 1.99 4.02 6.98 40 1.98 4.03 6.97 45 1.98 4.04 6.97 50 1.98 4.06 6.97 9.21 9.45 9.38 9.32 9.26 9.21 9.16 9.11 9.06 9.03 8.99 12.46 13.21 13.01 12.82 12.64 12.46 12.29 12.13 11.98 11.83 11.70 11.00 11.72 11.54 11.36 11.18 11.00 10.82 10.64 10.46 10.28 10.10 STARTER 5000 Bench pH Meter MERCK (Ref. 20°C) Temp °C 2.00 4.00 5 2.01 4.04 10 2.01 4.02 15 2.00 4.01 20 2.00 4.00 25 2.00 4.01 30 2.00 4.01 35 2.00 4.01 40 2.00 4.01 45 2.00 4.01 50 2.00 4.00 7.00 7.07 7.05 7.02 7.00 6.98 6.98 6.96 6.95 6.95 6.95 9.00 9.16 9.11 9.05 9.00 8.95 8.91 8.88 8.85 8.82 8.79 12.00 12.41 12.26 12.10 12.00 11.88 11.72 11.67 11.54 11.44 11.33 DIN (19267) (Ref. 25°C) Temp °C 1.09 4.65 6.79 5 1.08 4.67 6.87 10 1.09 4.66 6.84 15 1.09 4.66 6.82 20 1.09 4.65 6.80 25 1.09 4.65 6.79 30 1.10 4.65 6.78 35 1.10 4.65 6.77 40 1.10 4.66 6.76 45 1.10 4.67 6.76 50 1.11 4.68 6.76 9.23 9.43 9.37 9.32 9.27 9.23 9.18 9.13 9.09 9.04 9.00 12.75 13.63 13.37 13.16 12.96 12.75 12.61 12.45 12.29 12.09 11.98 JIS Z 8802 (Ref. 25°C) Temp °C 1.679 4.008 5 1.668 3.999 10 1.670 3.998 15 1.672 3.999 20 1.675 4.002 25 1.679 4.008 30 1.683 4.015 35 1.688 4.024 40 1.694 4.035 45 1.700 4.047 50 1.707 4.060 6.865 6.951 6.923 6.900 6.881 6.865 6.853 6.844 6.838 6.834 6.833 9.180 9.395 9.332 9.276 9.225 9.180 9.139 9.102 9.068 9.038 9.011 DIN(19266) / NIST (Ref. 25°C) Temp °C 1.680 4.008 6.865 5 1.668 4.004 6.950 10 1.670 4.001 6.922 15 1.672 4.001 6.900 20 1.676 4.003 6.880 25 1.680 4.008 6.865 30 1.685 4.015 6.853 35 1.691 4.026 6.845 40 1.697 4.036 6.837 45 1.704 4.049 6.834 50 1.712 4.064 6.833 9.183 9.392 9.331 9.277 9.228 9.183 9.144 9.110 9.076 9.046 9.018 12.454 13.207 13.003 12.810 12.627 12.454 12.289 12.133 11.984 11.841 11.705 EN-21 EN-22 JJG119 (Ref. 25°C, 0.001pH) Temp °C 1.680 4.003 6.864 5 1.669 3.999 6.949 10 1.671 3.996 6.921 15 1.673 3.996 6.898 20 1.676 3.998 6.879 25 1.680 4.003 6.864 30 1.684 4.010 6.852 35 1.688 4.019 6.844 40 1.694 4.029 6.838 45 1.700 4.042 6.834 50 1.706 4.055 6.833 STARTER 5000 Bench pH Meter 9.182 12460 9.391 13.210 9.330 13.011 9.276 12.820 9.226 12.637 9.182 12.460 9.142 12.292 9.105 12.130 9.072 11.975 9.042 11.828 9.015 11.697 STARTER 5000 Bench pH Meter EN-23 LIMITED WARRANTY Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus. This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages. As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for further details. TABLA DE CONTENIDO INTRODUCCIÓN ...............................................................1 1 1.1 1.2 2 Definición de advertencias y símbolos de señales ...................... 1 Precauciones de seguridad ............................................................ 2 INSTALACIÓN ...................................................................3 2.1 2.2 2.3 2.4 3 Contenido del paquete .................................................................... 3 Instalación del portaelectrodos independiente ............................ 4 Instalación del adaptador de corriente .......................................... 4 Conexión de los electrodos de pH ................................................. 5 PANTALLA E ÍCONOS ......................................................5 3.1 3.2 3.3 3.4 4 Encendido del medidor ................................................................... 5 Apagado del medidor ...................................................................... 6 Íconos................................................................................................ 6 Directorio del menú ......................................................................... 9 CALIBRACIÓN, MEDICIÓN E IMPRESIÓN .................... 10 4.1 4.1.1 4.1.2 4.2 4.2.1 4.2.2 4.3 4.4 5 Calibración...................................................................................... 10 Grupo de amortiguador .................................................................... 10 Calibración ........................................................................................11 Medición de muestras ................................................................... 12 Medición del pH ............................................................................... 12 Medición de mV y Rel.mV ............................................................... 12 Medición de temperatura .............................................................. 12 Impresión ........................................................................................ 13 CONFIGURACIÓN .......................................................... 14 5.1 Configuración del sistema ............................................................ 14 5.1.1 System setting (Configuración del sistema) ► Save Mode (Modo Guardar) ................................................................................ 14 5.1.2 System Setting (Configuración del sistema) ►Output (Salida) ......... 4 5.1.3 System Setting (Configuración del sistema) ► Restore Factory Setup (Restaurar configuración de fábrica) ..................................... 14 5.1.4 System Setting (Configuración del sistema) ► Service Password (Contraseña de servicio) .................................................................. 14 5.2 Configuración de medición........................................................... 15 5.2.1 Measurement Setting (Configuración de medición) ►Endpoint Mode (Modo de punto final) ....................................................................... 15 5.2.2 Measurement Setting (Configuración de medición) ►GLP Measurement Mode (Modo de medición GLP) ................................ 15 5.2.3 Measurement Setting (Configuración de medición) ►Continuous Measure (Medición continua) ........................................................... 15 5.3 Configuración de calibración ....................................................... 16 5.3.1 Calibration Setting (Configuración de calibración) ►Buffer Group (Grupo de amortiguador)................................................................. 16 5.3.2 Calibration Setting (Configuración de calibración) ►Calibration Reminder (Recordatorio de calibración) .......................................... 16 5.4 Registro de datos ........................................................................... 16 6 MANTENIMIENTO ........................................................... 17 6.1 6.2 6.3 6.4 7 DATOS TÉCNICOS ......................................................... 18 7.1 8 9 Mensaje de error ............................................................................ 17 Mantenimiento del medidor .......................................................... 17 Mantenimiento del electrodo ........................................................ 17 Información..................................................................................... 17 Especificaciones ............................................................................ 18 CONFORMIDAD NORMATIVA ........................................ 19 GRUPOS DE AMORTIGUADOR ..................................... 20 STARTER 5000 Medidor de pH de banco 1 ES-1 INTRODUCCIÓN Gracias por elegir este producto OHAUS. Lea este manual completamente antes de usar el medidor de pH de banco STARTER 5000 para evitar su operación incorrecta. El STARTER 5000 es un medidor de pH de 0.001 con una pantalla LCD táctil a color. Es una nueva generación de medidores de banco intuitivos y fáciles de usar para realizar mediciones confiables. Algunas de las tantas características actuales son:  Interfaz de usuario gráfica bilingüe  Punto final de tiempo automático o manual que puede ayudar a los usuarios a congelar el valor de lectura estable para atender diferentes requerimientos del cliente  8 grupos de amortiguador (US, EU, NIST y JJG) con 1 amortiguador autodefinido  Protección contra el ingreso de agua y polvo IP54 1.1 Definición de advertencias y símbolos de señales Las notas de seguridad están marcadas con palabras de señales y símbolos de advertencia. Estas muestran temas de seguridad y advertencias. Ignorar las notas de seguridad puede conducir a lesiones físicas, daños a los instrumentos, mal funcionamiento y resultados falsos. Palabras de señales ADVERTENCIA PRECAUCIÓN Atención Nota Para una situación peligrosa con riesgo medio que posiblemente resulte en lesiones o muerte si no se evita. Para una situación peligrosa de bajo riesgo, que resulta en daño al dispositivo o a la propiedad, pérdida de datos, o lesiones, si no se evita. Para información importante acerca del producto Para información útil acerca del producto Símbolos de advertencia Peligro general Riesgo de descarga eléctrica Peligro de explosión Peligro corrosivo Corriente alterna Corriente directa ES-2 1.2 STARTER 5000 Medidor de pH de banco Precauciones de seguridad PRECAUCIÓN Lea todas las advertencias de seguridad antes de instalar, hacer conexiones o dar servicio a este equipo. La falta de cumplimiento de estas advertencias puede resultar en lesiones personales o daños materiales. Conserve todas las instrucciones para referencia futura.           Verifique que el rango del voltaje de entrada del adaptador de corriente alterna y el tipo de enchufe sean compatibles con la alimentación de corriente alterna local. Verifique que el cordón eléctrico no represente un obstáculo potencial o riesgo de tropezarse. Use el equipo solamente en lugares secos. Seque de inmediato cualquier líquido que se derrame. El instrumento no es hermético al agua. Use solo accesorios y periféricos aprobados. Opere el equipo solo en las condiciones ambientales especificadas en estas instrucciones. Desconecte el equipo de la alimentación eléctrica antes de limpiarlo. No opere el equipo en ambientes peligrosos o inestables. El servicio debe proporcionarse solamente por personal autorizado. Si el equipo se utiliza de una manera no especificada por el fabricante,puede reducirse la prot ección provista por el equipo. ADVERTENCIA No use este instrumento en un ambiente sujeto a peligros de explosión. La caja del instrumento no es hermética al gas (peligro de explosión debido a formación de chispas; corrosión debido al ingreso de gases). PRECAUCIÓN Cuando use químicos y solventes, siga las instrucciones del fabricante y reglas de seguridad generales de laboratorio. STARTER Medidor de pH de banco 2 ES-3 INSTALACIÓN Desempaque con cuidado la caja. 2.1 Contenido del paquete El modelo ST5000-B (paquete básico) tiene los siguientes elementos: ST5000-B Medidor ST5000 Cubierta en uso de ST5000 Portaelectrodos independiente Unidad USB de 4 GB Juego de polvo amortiguador (4.01; 7.00; 10.00) Alimentación de energía Unidades 1 1 1 juego 1 1 juego 1 juego Además del contenido del ST5000-B, el paquete del ST5000-F también incluye lo siguiente: Electrodo de pH rellenable de vidrio 3 en 1 ST350 1 Cada polvo amortiguador de pH debe disolverse en 250 ml de agua pura o agua desionizada en un matraz volumétrico. También puede ordenar solución amortiguadora embotellada de Ohaus. Para más electrodos, soluciones, accesorios y partes de repuesto, comuníquese a Ohaus para más detalles. Modelo ST350 ST310 ST210 ST320 ST260 ST230 STSURF STMICRO5 STMICRO8 STTEMP30 STORP1 STORP2 Descripción Electrodo de pH rellenable de vidrio 3 en 1 Electrodo de pH rellenable de plástico 3 en 1 Electrodo de pH rellenable de plástico 2 en 1 Electrodo de pH de gel de plástico 3 en 1 (no se necesita llenar) Electrodo de pH de puente de sal doble de vidrio 2 en 1 Electrodo de pH de muestra fangosa de vidrio 2 en 1 Electrodo de pH de superficie plana 2 en 1 Electrodo de pH de muestra micro de vidrio 2 en 1 Electrodo de pH de muestra micro de vidrio 2 en 1 Electrodo de temperatura Electrodo ORP de plástico de gel Electrodo ORP de vidrio rellenable Juego de polvo amortiguador (4.01; 7.00; 10.0) Amortiguador pH 1.68 250 ml Amortiguador pH 4.01 250 ml Amortiguador pH 7.00 250 ml Amortiguador pH 10.01 250 ml Amortiguador pH 12.45 250 ml Electrólito de referencia de electrodo de pH Solución de protección de electrodo de pH Portaelectrodos independiente Cubierta en uso de ST5000 Adaptador de cable SP RS232 F-M SF40A (para ST5000) Número de parte 30129354 83033965 83033966 83033967 30129357 83033968 30129470 30087566 30087569 83033970 30038555 30038553 83033971 30100424 30100425 30100427 30100429 30100440 30059255 30059256 30058733 30129897 30059316 ES-4 2.2 STARTER 5000 Medidor de pH de banco Instalación del portaelectrodos independiente Instale el brazo del electrodo en la base: Después de ajustar la perilla de apriete hasta cierto punto, el brazo superior puede moverse hacia arriba y abajo. Cuando el electrodo de pH se coloca en el brazo y se realiza la calibración o medición, el frasco de almacenamiento del electrodo de pH puede colocarse en el orificio de la base para fácil almacenamiento: 2.3 Instalación del adaptador de corriente Inserte el enchufe del adaptador en la ranura del adaptador. PRECAUCIÓN Solamente use un enchufe de adaptador que sea compatible con la toma de corriente de la alimentación eléctrica. Asegúrese de que el enchufe del adaptador esté apoyado completamente sobre la ranura del adaptador antes de conectarlo en la toma de corriente de la alimentación eléctrica. STARTER Medidor de pH de banco 2.4 ES-5 Conexión de los electrodos de pH Hay dos conectores para el electrodo. Conector “pH” (BNC); conector “TEMP.” (Cinch). Para el electrodo ST350 3 en 1, conecte los conectores BNC y Cinch. Para los electrodos de pH 2 en 1 (por ejemplo, el ST230), solamente conecte el BNC. 3 PANTALLA E ÍCONOS 3.1 Encendido del medidor El STARTER 5000 es un medidor de banco de pH 0.001 de pantalla táctil a color. No tiene ningún botón. Después de conectar la corriente, el medidor se encenderá; si tiene protector de pantalla, toque la pantalla para encenderla. Nota: si el cliente no ha trabajado con el medidor durante más de 6 horas, el medidor se apagará automáticamente. En este caso ya no podrá encender el medidor con solo tocar la pantalla. Es necesario volver a conectar la corriente para encender el medidor. Seleccione el idioma y presione Interfaz principal para entrar en la interfaz principal. ES-6 3.2 STARTER 5000 Medidor de pH de banco Apagado del medidor Para apagar el medidor, toque el botón de apagado en la parte superior izquierda de la interfaz principal para entrar en apagar interfaz. Luego toque medidor, o toque para apagar el para volver a la interfaz principal. Apagar la interfaz 3.3 Íconos Identificación de usuario Identificación de la muestra Ícono de información Fecha Hora Modo de medición de pH. Alterna entre el modo mV y Rel.mV o Ícono de señal; cuando destella significa que el medidor está recibiendo señal (en medición o en calibración). Íconos de punto final, punto final automático, punto final manual y punto final de tiempo. Cuando está en punto final manual, si la señal es estable, el medidor mostrará el ícono de señal estable. Toque el ícono de detener manual. Ícono de señal estable. Ícono de temperatura. para alcanzar el punto final STARTER 5000 Medidor de pH de banco ES-7 Área de toque de información, medición y calibración El área de información también es el acceso rápido para entrar en la interfaz de configuración: Toque esta área para entrar en la interfaz para configurar el modo de punto "Endpoint:" final. Los modos disponibles son punto final automático (Auto), Manual y de tiempo (Time). Se muestra el modo de punto final actual. Toque esta área para entrar en la interfaz para seleccionar, configurar o "Sensor ID:" cambiar la identificación del sensor. Se muestra la identificación del sensor actual. "Sensor Cond:" Toque esta área para entrar en la interfaz de datos de calibración del sensor. Se muestra el ícono de rendimiento del sensor actual (según los datos de calibración). Hay 3 íconos de condición de sensor: Pendiente: más del 95% y desfase: ± (0-15) mV La condición del electrodo es buena Pendiente: 90-95% o desfase: ± (15-35) mV La condición del electrodo es aceptable Pendiente: menos del 90% o desfase: ± (35-60) mV La condición del electrodo no es buena o necesita limpieza Ícono de recordatorio de calibración. Ajuste la hora en que desee que el medidor le avise para hacer una calibración desde la última calibración. Cuando se llegue a la hora, se mostrará este ícono y destellará para recordarle que es momento de realizar una calibración. “Buffers:” Esta área muestra el grupo de amortiguador actual; toque esta área para entrar a la interfaz y seleccionar el grupo de amortiguador. Ícono de inicio. Toque este ícono y el medidor comenzará la medición; el ícono de señal destellará. El área del ícono de inicio mostrará el ícono de detener Ícono de calibración. Toque este ícono para iniciar la calibración. Ícono de guardar. Después de guardar el resultado de la medición actual (se alcanzó el punto final), el color del ícono cambiará de negro a gris claro para indicar que la lectura ya se guardó y no puede guardarse dos veces. Ícono de salida. Envíe datos a un dispositivo externo tal como una impresora. Si se selecciona “Shut off connection” (Apagar conexión) en System Setting (Configuración del sistema) ►Output (Salida), se mostrará el ícono de guardar posición. . De lo contrario, el ícono de salida se mostrará en la misma ES-8 STARTER 5000 Medidor de pH de banco Volver a la interfaz principal Volver al menú anterior Entrar, confirmar o aplicar la configuración actual, terminar la configuración Editar Retroceder página Avanzar página Eliminar Incrementar (luz) Reducir (luz) Pantalla de gráfico (para los datos de medición o calibración) Guardar Salida a dispositivo externo (unidad USB, impresora) Teclado completo Siguiente punto de calibración Punto de calibración actual nuevamente Fin de calibración STARTER 5000 Medidor de pH de banco 3.4 Directorio del menú Los menús del primer y segundo directorio para el STARTER 5000: Fecha Hora Modo Guardar Salida Configuración del sistema Brillo de la pantalla Tasa de baudios Protección de pantalla Restaurar configuración de fábrica Información del instrumento Contraseña de servicio Resolución Entrada de temperatura MTC Modo pH/mV/Rel.mV Configuración de medición Medición continua Modo Punto final Configuración de identificación Modo Medición GLP Configuración de calibración Datos Grupo de amortiguador Recordatorio de calibración Datos de medición Datos de calibración ES-9 ES-10 4 STARTER 5000 Medidor de pH de banco CALIBRACIÓN, MEDICIÓN E IMPRESIÓN La pantalla del medidor se apagará automáticamente de acuerdo con la hora configurada en el menú System Setting (Configuración del sistema) ►Screen Protection (Protección de pantalla) ►LCD Auto-off. (Apagado automático de LCD) Para volver a encender la pantalla, toque la pantalla (vea 3.1) El proceso de medición de pH deberá seguir este procedimiento: a) Enjuagar el electrodo de pH y conectarlo al medidor b) Preparar el amortiguador y calibrar el electrodo de pH c) Preparar la muestra d) Realizar la medición de pH de la muestra e) Registrar los datos de la medición f) Enjuagar el electrodo de pH y almacenarlo Preparación del electrodo del pH: el electrodo de pH deberá enjuagarse con agua pura antes y después de usarse. Revise si el electrodo presenta algún daño físico. (Tenga cuidado con el bulbo de vidrio.) El electrodo de pH debe almacenarse en el frasco de almacenamiento. Después de que el electrodo de pH se haya puesto en la muestra o solución amortiguadora, agite varios segundos y luego espere de 30 a 60 segundos para que la señal sea estable, y entonces presione el botón para operar (calibración o medición). ADVERTENCIA No opere el equipo en ambientes peligrosos. El equipo no tiene protección contra explosión. PRECAUCIÓN Cuando use químicos y solventes, siga las instrucciones del fabricante del químico y las reglas de seguridad generales de laboratorio. 4.1 Calibración 4.1.1 Grupo de amortiguador El STARTER 5000 puede realizar calibraciones de 1 a 5 puntos para grupos predeterminados de amortiguador. Para el grupo de amortiguador autodefinido, puede realizar calibración de 1 a 9 puntos. Antes de la calibración, asegúrese de que el grupo de amortiguador actual es el grupo deseado. Cuando use un nuevo electrodo de pH, deberá realizarse una calibración antes de cualquier medición. Si un electrodo de pH no se usa durante varios días, deberá realizarse una calibración antes del siguiente uso. La calibración debe mostrar el valor correcto de pH cuando el medidor reciba la señal del valor de mV del electrodo de pH. Pendiente: el coeficiente lineal entre mV y pH de acuerdo con el valor teórico (por ejemplo, -59.16 mV/pH a 25 °C significa una pendiente del 100%); Desfase: el valor mV cuando el valor del pH es 7.00. (El valor teórico es 0 mV.) STARTER 5000 Medidor de pH de banco ES-11 El STARTER 5000 presentará un ícono de condición de electrodo para mostrar el desempeño del electrodo de pH de acuerdo con los datos de calibración. Normalmente, si vemos un ícono de cara triste después de una calibración adecuada, es posible que necesite reemplazar el electrodo de pH por uno nuevo. El STARTER 5000 se corrige automáticamente para la dependencia de temperatura de los valores de pH del amortiguador. Por ejemplo, para el grupo de amortiguador estándar de Estados Unidos "1.68, 4.01, 7.00, 10.01", consulte la siguiente tabla: 5 °C 1.67 4.01 7.09 10.25 10 °C 1.67 4.00 7.06 10.18 15 °C 1.67 4.00 7.04 10.12 20 °C 1.68 4.00 7.02 10.06 25 °C 1.68 4.01 7.00 10.01 30 °C 1.68 4.01 6.99 9.97 35 °C 1.69 4.02 6.98 9.93 40 °C 1.69 4.03 6.97 9.89 45 °C 1.70 4.05 6.97 9.86 50 °C 1.71 4.06 6.96 9.83 Por ejemplo, con el amortiguador de calibración 10.01 a 15 °C, el resultado de calibración deberá ser pH 10.12. 4.1.2 Calibración Después de enjuagar el electrodo de pH, coloque el electrodo en el primer amortiguador, agite durante 10 segundos y luego presione para iniciar la calibración. Cuando está en calibración, el ícono de señal destella. Cuando se ha alcanzado el punto final, aparece mientras desaparece y la calibración del punto actual es finalizada. Ahora hay 3 opciones: Salir de la calibración y volver a la interfaz principal Terminar la calibración del punto actual e ir a la interfaz del resultado de calibración Hacer la calibración del punto siguiente Nota: durante el proceso de calibración, el ST5000 siempre está en Auto Endpoint Mode (Modo de punto final automático). Nota: se recomienda usar un electrodo de pH 3 en 1 que incluya el sensor de temperatura o un electrodo de temperatura aparte. Cuando use el modo MTC, ingrese el valor de temperatura correcto y mantenga todas las soluciones amortiguadoras y de muestra a la temperatura establecida. ES-12 STARTER 5000 Medidor de pH de banco 4.2 Medición de muestras 4.2.1 Medición del pH Asegúrese de que el electrodo se haya calibrado antes de la medición. Después de enjuagar,  Coloque el electrodo de pH en la muestra.  Presione para comenzar la medición de pH; cuando se haya alcanzado el punto final, la medición actual está terminada. Nota: si la señal no es estable todo el tiempo, se recomienda usar el modo de punto final manual para evitar la repetibilidad errónea. 4.2.2 Medición de mV y Rel.mV Presione para cambiar al modo o ; siga el mismo procedimiento que para la medición del pH para la medición de mV o Rel.mV. Si el medidor está conectado con un electrodo ORP (potencial de oxidación reducción), el resultado de mV es el valor del valor “potencial de oxidación-reducción”, también llamado valor Redox. 4.3 Medición de temperatura Para mayor precisión, recomendamos usar un electrodo de temperatura integrado o separado.  Si se usa un electrodo de temperatura, se muestran ATC y la temperatura de la muestra.  Si el medidor no detecta un electrodo de temperatura, automáticamente cambia al modo de compensación de temperatura manual y aparece MTC. Se debe establecer la temperatura MTC. Nota: el STARTER 5000 acepta sensor de temperatura NTC de 30 kΩ. Para ATC (compensación de temperatura automática) o MTC (compensación de temperatura manual), la compensación de temperatura solo se corrige para el cambio en la salida del electrodo, no para el cambio en la solución real. Eso significa que el medidor corrige la señal (mV) de los electrodos de pH para obtener un valor de pH más preciso de acuerdo con la temperatura real. STARTER 5000 Medidor de pH de banco 4.4 ES-13 Impresión El STARTER 5000 puede imprimir los datos de medición o calibración a través del puerto RS232. Después de que la impresora (por ejemplo, OHAUS SF-40A) se conecte al medidor correctamente, realice las siguientes configuraciones (vea 5.1): ordene el cable de conexión para ST5000 (N/P 30059316) para conectar SF40A y ST5000. System Setting (Configuración del sistema) ► Output (Salida) , seleccione Printer (Impresora) System Setting (Configuración del sistema) ► Baud Rate (Tasa de baudios), seleccione 9600. El formato para la impresión de la medición es: Fecha Hora Usuario Id. de la muestra 6-13-2015 14:20:22 Usuario01 Muestra02 Id. sensor Modo punto final MTC/ATC Sensor02 Manual ATC Temperatura Resultado 25.3 ºC 6.999pH El formato para la impresión de calibración es: Datos de calibración Id. de sensor Número de serie Sensor02 14060012 Fecha cal. Hora cal. ATC/MTC 01-01-2015 14:20:33 ATC Amortig. mV Desfase Pendiente 4.00pH 176.94mV 7.00pH 0.05mV 0.05mV 98.22% Cuando se realizan mediciones continuas, el formato para imprimir los datos de calibración es: Tiempo Medición Registro Id. de sensor Número de serie Sensor02 14060012 Fecha Hora Temperatura 01-01-2015 14:20:25 23.5 ºC Fecha Hora Temperatura 01-01-2015 14:20:30 23.5 ºC Fecha Hora Temperatura 01-01-2015 14:20:35 23.5 ºC Medición 6.998pH Medición 6.999pH Medición 7.000pH ES-14 STARTER 5000 Medidor de pH de banco 5 CONFIGURACIÓN 5.1 Configuración del sistema La configuración del sistema se usa para establecer parámetros del medidor; la información del medidor también podría encontrarse en este. Vea la sección 3.4 para referencia. 5.1.1 System setting (Configuración del sistema) ►Save Mode (Modo Guardar) El modo Guardar tiene 2 opciones, uno es “Auto Save Endpoint Reading” (Guardar automáticamente lectura del punto final) y el otro es “Manual Save Endpoint Reading” (Guardar manualmente la lectura del punto final). “Auto Save Endpoint Reading” significa que cuando la medición alcanza el punto final (punto final automático, manual u hora), el medidor guardará automáticamente la lectura del punto final. El ícono Guardar cambiará de negro a gris claro solo puede guardarse una vez. ; una lectura del punto final 5.1.2 System Setting (Configuración del sistema) ►Output (Salida) Cuando se selecciona apagar Guardar 5.1.3 la conexión, la interfaz principal mostrará el , o mostrará el ícono Salida ícono . System Setting (Configuración del sistema) ► Restore Factory Setup (Restaurar configuración de fábrica) La restauración de la configuración de fábrica incluye lo siguiente: 1. Apagar el modo GLP 2. Apagar el modo Guardar automáticamente 3. Apagar la medición continua 4. Apagar la conexión (salida) 5. Apagar el recordatorio de calibración 6. La resolución es: 0.001pH, 0.01mV 7. Modo de medición: pH: 8. Se eliminará el grupo de amortiguador autodefinido 9. Tasa de baudios: 9600 10. Brillo de la pantalla: 70% 11. Apagar la protección de pantalla 12. MTC: 25 ºC 13. Punto final: Automático 14. Punto final de tiempo: 10 segundos 15. Eliminar todos los datos de calibración 5.1.4 System Setting (Configuración del sistema) ► Service Password (Contraseña de servicio) No se usa. STARTER 5000 Medidor de pH de banco ES-15 5.2 Configuración de medición 5.2.1 Measurement Setting (Configuración de medición) ►Endpoint Mode (Modo de punto final) Punto final automático En este modo, el medidor alcanzará el punto final y bloqueará la lectura automáticamente de acuerdo con los siguientes criterios de estabilidad: Resolución 0.1pH: la señal de la entrada del sensor puede no cambiar más de 0.3 mV en 6 segundos. Resolución 0.01pH: la señal de la entrada del sensor puede no cambiar más de 0.1 mV en 6 segundos. Resolución 0.001pH: la señal de la entrada del sensor puede no cambiar más de 0.03 mV en 6 segundos. Punto final de tiempo En este modo, el usuario necesita ingresar el tiempo del punto final, por ejemplo 15 segundos. Esto significa que después de iniciar la medición durante 15 segundos, el medidor alcanzará el punto final. Punto final manual En este modo, el usuario necesita decidir cuándo es estable la señal y luego presionar el ícono para detener 5.2.2 para alcanzar el punto final manual. Measurement Setting (Configuración de Measurement Mode (Modo de medición GLP) medición) ►GLP El modo GLP requiere que cada medición tenga la identificación del usuario y la identificación de la muestra. Si no hay identificación del usuario e identificación de la muestra, el medidor mostrará la interfaz para permitir que el usuario ingrese los datos requeridos. 5.2.3 Measurement Setting (Configuración de medición) ►Continuous Measure (Medición continua) Durante la medición continua, el usuario necesita ingresar el intervalo de tiempo para obtener cada resultado de lectura. A continuación se muestra un Gráfico de medición continua. El eje Y es el valor de pH. El eje x es el número del resultado de la medición. ES-16 STARTER 5000 Medidor de pH de banco 5.3 Configuración de calibración 5.3.1 Calibration Setting (Configuración de calibración) ►Buffer Group (Grupo de amortiguador) Antes de la calibración, asegúrese de que el grupo de amortiguador actual es el correcto para sus soluciones amortiguadoras. El usuario puede seleccionar entre 8 grupos de amortiguador definidos. El usuario también puede establecer grupos de amortiguador autodefinidos. La diferencia de 2 puntos de calibración adyacentes deberá ser mayor de 0.99pH. 5.3.2 Calibration Setting (Configuración de calibración) ►Calibration Reminder (Recordatorio de calibración) Ingrese a esta configuración para ajustar el tiempo, desde la última calibración, para determinar después de cuántas horas el medidor debe recordar realizar la calibración nuevamente. El medidor mostrará el ícono de recordatorio de calibración usuario realizar la calibración. 5.4 y destellará para recordar al Registro de datos Toque Datos para entrar en la interfaz del registro de datos, incluyendo datos de medición y datos de calibración. Aquípuede eliminar todos los datos de medición. La única forma de eliminar datos de calibración es restaurando la configuración de fábrica. Los datos pueden enviarse a una unidad USB (≤ 4 GB). STARTER 5000 Medidor de pH de banco 6 MANTENIMIENTO 6.1 Mensaje de error ES-17 Cuando el valor ingresado o resultado de la medición esté fuera de rango o los datos de calibración no sean suficientemente buenos, el medidor mostrará información de error. Si el error no puede resolverse mediante el mensaje de error mostrado, comuníquese con el representante de ventas de Ohaus para mayor asistencia. 6.2 Mantenimiento del medidor Los instrumentos de la serie STARTER 5000 no requieren ningún mantenimiento excepto la limpieza ocasional con una tela humedecida. Atención: siga estas precauciones de seguridad cuando limpie el instrumento.  Desconecte el instrumento de la alimentación eléctrica  No abra el instrumento  No permita que los líquidos penetren en el instrumento Atención: la caja está hecha de acronitrilo butadieno estireno (ABS). Este material es atacado por algunos solventes orgánicos como el tolueno, el xileno y el metil etil cetona (MEK). Cualquier derrame deberá limpiarse inmediatamente. 6.3 Mantenimiento del electrodo Asegúrese de que el electrodo esté lleno con solución de electrólito. Siempre almacene el electrodo de acuerdo con el manual de instrucciones del electrodo y no permita que se seque. Si la respuesta del electrodo llega a ser lenta o la pendiente no es suficientemente buena, intente lo siguiente: o Remoje el electrodo en HCl 0.1 M durante más de 8 horas. o Para contaminación con grasa o aceite, desengrase la membrana con lana de algodón empapada en acetona o solución de jabón. Después del tratamiento del electrodo, deberá realizarse una nueva calibración. Si la pendiente del electrodo aún no es buena, el electrodo podría necesitar reemplazarse. 6.4 Información Toque el ícono de información para entrar en la interfaz de información. El usuario puede ingresar aquíinformación relacionada al tocar el área que tiene " " en la parte inferior izquierda, tal como fecha de compra del sensor y otra. ES-18 STARTER 5000 Medidor de pH de banco 7 DATOS TÉCNICOS 7.1 Especificaciones Los datos técnicos son válidos en las siguientes condiciones ambientales:  Para uso exclusivo en interiores  Altitud: hasta 2000 m  Rango de temperatura especificado: 5 °C a 40 °C  Humedad: humedad relativa máxima del 80 % para temperaturas hasta de 30 °C, disminuyendo linealmente al 50% de humedad relativa a 40 °C  Fluctuaciones de voltaje de la alimentación principal: hasta ± 10% del voltaje nominal  Categoría de instalación II  Grado de contaminación: 2  La operabilidad se garantiza a temperaturas ambiente entre 5 °C a 40 °C Modelo STARTER 5000 Rango de medición -2.000…20.000 pH –2000.00…2000.00 mV -30 °C…130 °C Resolución 0.1/0.01/0.001 pH 1/ 0.1/ 0.01 mV 0.1 °C Límites de error ± 0.002 pH ± 0.03% FS mV ± 0.1 °C Calibración 1 a 9 puntos 8 grupos de amortiguador predefinidos más 1 grupo de amortiguador autodefinido 1000 mediciones 10*10 datos de calibración Memoria Alimentación de energía Entrada de 100-240 V ~0.3 A, 9V 1.12 A de salida Adaptador de CA de Tamaño/peso Aproximadamente 210 x 170 x 80 mm / 0.65 kg Dimensión/peso de envío Aproximadamente 372 x 310 x 185 mm / 2.30 kg Pantalla Pantalla táctil de cristal líquido con iluminación de fondo Entrada BNC, impedancia > 3 x 10e+12 Ω Cinch, NTC 30 kΩ Compensación de temperatura ATC y MTC Caja ABS STARTER 5000 Medidor de pH de banco 8 ES-19 CONFORMIDAD NORMATIVA Este producto está en conformidad con la Directiva EMC 2004/108/EC y con la Directiva para Bajo Voltaje 2006/95/EC. La Declaración de conformidad está disponible en línea en europe.ohaus.com/europe/en/home/support/compliance.aspx. De conformidad con la Directiva Europea 2002/96/EC sobre Residuos de Equipos Eléctricos y Electrónicos (WEEE), este dispositivo no puede desecharse con la basura doméstica. Esto también es aplicable para países fuera de la UE, según sus requerimientos específicos. La Directiva para baterías 2006/66/EC introduce nuevos requisitos a partir de septiembre de 2008 acerca de la remoción de las baterías de equipos de desecho en los países miembros de la Unión Europea. Para cumplir con esta Directiva, este dispositivo ha sido diseñado para remoción segura de baterías al final de su vida útil a través de un centro de tratamiento de desechos. Deseche este producto de acuerdo con las regulaciones locales en el punto de recolección especificado para equipos eléctricos y electrónicos. Si tiene alguna pregunta, comuníquese con la autoridad responsable o con el distribuidor a quien compró este dispositivo. En caso que este dispositivo sea transferido a otras partes (para uso privado o profesional), también deberá mencionarse el contenido de esta regulación. Las instrucciones de desecho en Europa están disponibles en línea en europe.ohaus.com/europe/en/home/support/weee.aspx. Gracias por su contribución a la protección ambiental. Nota de la FCC: Este equipo ha sido probado y se encontró que cumple con los límites para un dispositivo digital clase A, consecuente con la Parte 15 de las Reglas de la FCC. Estos límites están diseñados para proporcionar una protección razonable contra interferencia dañina cuando el equipo es operado en un ambiente comercial. Este equipo genera, usa y puede irradiar frecuencias de radio y, si no es instalado y utilizado de acuerdo con el manual de instrucciones, puede causar interferencia dañina a las radiocomunicaciones. Es probable que la operación de este equipo en un área residencial cause interferencia dañina, en cuyo caso se le exigirá al usuario que corrija la interferencia con gastos a su cargo. Nota de Industry Canada Este aparato digital Clase A cumple con la norma canadiense ICES-001. Certificación ISO 9001 En 1994, Ohaus Corporation de Estados Unidos recibió la certificación ISO 9001 de parte de Bureau Veritas Quality International (BVQI), lo cual confirmó que el sistema de gestión de calidad de Ohaus cumple con los requerimientos de la norma ISO 9001. El 21 de junio de 2012, Ohaus Corporation de Estados Unidos recibió la recertificación de la norma ISO 9001:2008. ES-20 9 STARTER 5000 Medidor de pH de banco GRUPOS DE AMORTIGUADOR El STARTER 5000 corrige automáticamente la dependencia de temperatura del valor de pH del grupo de amortiguador dado en las siguientes tablas. Seleccione el grupo de amortiguador que desea antes de la calibración. JJG119 (Ref. 25 °C, 0.01pH) Temp. °C 1.68 4.00 5 1.67 4.00 10 1.67 4.00 15 1.67 4.00 20 1.68 4.00 25 1.68 4.00 30 1.68 4.01 35 1.69 4.02 40 1.69 4.03 45 1.70 4.04 50 1.71 4.06 6.86 6.95 6.92 6.90 6.88 6.86 6.85 6.84 6.84 6.83 6.83 9.18 9.39 9.33 9.28 9.23 9.18 9.14 9.11 9.07 9.04 9.02 Estándar de los EE. UU. (Ref. 25 °C) Temp. °C 1.68 4.01 7.00 5 1.67 4.01 7.09 10 1.67 4.00 7.06 15 1.67 4.00 7.04 20 1.68 4.00 7.02 25 1.68 4.01 7.00 30 1.68 4.01 6.99 35 1.69 4.02 6.98 40 1.69 4.03 6.97 45 1.70 4.05 6.97 50 1.71 4.06 6.96 10.01 10.25 10.18 10.12 10.06 10.01 9.97 9.93 9.89 9.86 9.83 Estándar de Europa (Ref. 25 °C) Temp. °C 2.00 4.01 7.00 5 2.02 4.01 7.09 10 2.01 4.00 7.06 15 2.00 4.00 7.04 20 2.00 4.00 7.02 25 2.00 4.01 7.00 30 1.99 4.01 6.99 35 1.99 4.02 6.98 40 1.98 4.03 6.97 45 1.98 4.04 6.97 50 1.98 4.06 6.97 9.21 9.45 9.38 9.32 9.26 9.21 9.16 9.11 9.06 9.03 8.99 12.46 13.21 13.01 12.82 12.64 12.46 12.29 12.13 11.98 11.83 11.70 11.00 11.72 11.54 11.36 11.18 11.00 10.82 10.64 10.46 10.28 10.10 STARTER 5000 Medidor de pH de banco MERCK (Ref. 20 °C) Temp. °C 2.00 4.00 5 2.01 4.04 10 2.01 4.02 15 2.00 4.01 20 2.00 4.00 25 2.00 4.01 30 2.00 4.01 35 2.00 4.01 40 2.00 4.01 45 2.00 4.01 50 2.00 4.00 7.00 7.07 7.05 7.02 7.00 6.98 6.98 6.96 6.95 6.95 6.95 9.00 9.16 9.11 9.05 9.00 8.95 8.91 8.88 8.85 8.82 8.79 12.00 12.41 12.26 12.10 12.00 11.88 11.72 11.67 11.54 11.44 11.33 DIN (19267) (Ref. 25 °C) Temp. °C 1.09 4.65 6.79 5 1.08 4.67 6.87 10 1.09 4.66 6.84 15 1.09 4.66 6.82 20 1.09 4.65 6.80 25 1.09 4.65 6.79 30 1.10 4.65 6.78 35 1.10 4.65 6.77 40 1.10 4.66 6.76 45 1.10 4.67 6.76 50 1.11 4.68 6.76 9.23 9.43 9.37 9.32 9.27 9.23 9.18 9.13 9.09 9.04 9.00 12.75 13.63 13.37 13.16 12.96 12.75 12.61 12.45 12.29 12.09 11.98 JIS Z 8802 (Ref. 25 °C) Temp. °C 1.679 4.008 5 1.668 3.999 10 1.670 3.998 15 1.672 3.999 20 1.675 4.002 25 1.679 4.008 30 1.683 4.015 35 1.688 4.024 40 1.694 4.035 45 1.700 4.047 50 1.707 4.060 6.865 6.951 6.923 6.900 6.881 6.865 6.853 6.844 6.838 6.834 6.833 9.180 9.395 9.332 9.276 9.225 9.180 9.139 9.102 9.068 9.038 9.011 DIN(19266) / NIST (Ref. 25 °C) Temp. °C 1.680 4.008 6.865 5 1.668 4.004 6.950 10 1.670 4.001 6.922 15 1.672 4.001 6.900 20 1.676 4.003 6.880 25 1.680 4.008 6.865 30 1.685 4.015 6.853 35 1.691 4.026 6.845 40 1.697 4.036 6.837 45 1.704 4.049 6.834 50 1.712 4.064 6.833 9.183 9.392 9.331 9.277 9.228 9.183 9.144 9.110 9.076 9.046 9.018 12.454 13.207 13.003 12.810 12.627 12.454 12.289 12.133 11.984 11.841 11.705 ES-21 ES-22 JJG119 (Ref. 25 °C, 0.001pH) Temp. °C 1.680 4.003 6.864 5 1.669 3.999 6.949 10 1.671 3.996 6.921 15 1.673 3.996 6.898 20 1.676 3.998 6.879 25 1.680 4.003 6.864 30 1.684 4.010 6.852 35 1.688 4.019 6.844 40 1.694 4.029 6.838 45 1.700 4.042 6.834 50 1.706 4.055 6.833 STARTER 5000 Medidor de pH de banco 9.182 9.391 9.330 9.276 9.226 9.182 9.142 9.105 9.072 9.042 9.015 12460 13.210 13.011 12.820 12.637 12.460 12.292 12.130 11.975 11.828 11.697 STARTER 5000 Medidor de pH de banco ES-23 GARANTÍA LIMITADA Los productos de Ohaus están garantizados contra defectos de materiales y fabricación desde la fecha de entrega y durante el tiempo que dure la garantía. Durante el periodo de garantía Ohaus reparará, o, a su discreción, reemplazará cualquier componente(s) que se compruebe que está defectuoso, sin costo alguno, con la condición de que el producto sea devuelto, flete pagado, a Ohaus. Esta garantía no es válida si el producto ha sufrido daños por accidente o mal uso, ha sido expuesto a materiales radioactivos o corrosivos, presenta materiales extraños penetrando al interior del producto, o si ha sido objeto de modificaciones por parte de personas que no sean de Ohaus. En el lugar de una tarjeta de registro de garantía devuelta debidamente, el periodo de garantía comenzará en la fecha de envío al distribuidor autorizado. Ohaus Corporation no otorga ninguna otra garantía expresa o implícita. Ohaus Corporation no se hace responsable de cualquier daño resultante. Puesto que la legislación sobre garantías difiere de un estado a otro asícomo de un país a otro, comuníquese con su distribuidor local de Ohaus para mayor información. TABLE DES MATIÈRES INTRODUCTION................................................................ 1 1 1.1 1.2 2 Définition des symboles et des indicateurs d'avertissements ... 1 Précautions de sécurité .................................................................. 2 INSTALLATION ................................................................. 3 Contenu de l’emballage ................................................................... 3 Installation du porte-électrode autonome ..................................... 4 Installation de l'adaptateur d'alimentation .................................... 4 Connexion des électrodes de pH ................................................... 5 2.1 2.2 2.3 2.4 3 ICÔNES ET AFFICHAGE .................................................. 5 Mise sous tension de l’appareil ...................................................... 5 Mise hors tension de l'appareil....................................................... 6 Icônes ................................................................................................ 6 Répertoire du menu ......................................................................... 9 3.1 3.2 3.3 3.4 4 ÉTALONNAGE, MESURE ET IMPRESSION .................. 10 4.1 4.1.1 4.1.2 4.2 4.2.1 4.2.2 4.3 4.4 5 Étalonnage ...................................................................................... 10 Groupe de tampons ......................................................................... 10 Réalisation d'un étalonnage..............................................................11 Mesure d'un échantillon ................................................................ 12 Mesure du pH .................................................................................. 12 Mesure en mV et Rel.mV ................................................................. 12 Mesure de température ................................................................. 12 Impression ...................................................................................... 13 CONFIGURATION ........................................................... 14 5.1 5.1.1 5.1.2 5.1.3 5.1.4 5.2 5.2.1 5.2.2 5.2.3 5.3 5.3.1 5.3.2 5.4 6 Paramétrage système .................................................................... 14 Paramétrage système►Mode Enregistrement................................ 14 Paramétrage système►Sortie ......................................................... 14 Paramétrage système►Restauration des paramètres par défaut .. 14 Paramétrage système► Mot de passe de service .......................... 14 Paramétrage des mesures ............................................................ 15 Paramétrage des mesures ►Mode de résultat final ....................... 15 Paramétrage des mesures ►Mode GLP de mesure ...................... 15 Paramétrage des mesures ►Mesure en continu ............................ 15 Paramétrage de l'étalonnage ........................................................ 16 Paramétrage de l'étalonnage ►Groupe de tampons ...................... 16 Paramétrage de l'étalonnage ►Rappel de l'étalonnage ................. 16 Journal des données ..................................................................... 16 MAINTENANCE............................................................... 17 6.1 Message d’erreur ........................................................................... 17 6.2 6.3 6.4 7 DONNÉES TECHNIQUES ............................................... 18 7.1 8 9 Maintenance de l'appareil de mesure .......................................... 17 Maintenance de l'électrode ........................................................... 17 Informations ................................................................................... 17 Spécifications ................................................................................. 18 CONFORMITÉ ................................................................. 19 GROUPES DE TAMPONS .............................................. 20 PH-mètre de laboratoire STARTER 5000 1 FR-1 INTRODUCTION Merci d'avoir choisi ce produit OHAUS. Lire le manuel complètement avant d'utiliser le pH-mètre de laboratoire STARTER 5000 afin d'éviter une exploitation inadéquate. Le STARTER 5000 est un pH-mètre de laboratoire (0,001) avec un écran ACL tactile couleur. Il s'agit d'une nouvelle génération d'appareils intuitifs et faciles d'utilisation permettant des mesures fiables. Suivent ici quelques-unes des nombreuses fonctions existantes :  Interface utilisateur graphique bilingue  Indicateur temporel de résultats automatiques et manuels pouvant aider les utilisateurs à figer un résultat stable afin de répondre aux différents impératifs de la clientèle  8 groupes de tampons (US, EU, NIST et JJG) avec un tampon auto-défini  Protection IP54 contre l'eau et la poussière 1.1 Définition des symboles et des indicateurs d'avertissements Les remarques de sécurité sont signalées par des mots indicateurs et par des symboles d'avertissement. Elles présentent les problèmes et les avertissements de sécurité. L'inobservation des remarques de sécurité peut conduire à des accidents, à l'endommagement de l'instrument, à des dysfonctionnements et à des résultats erronés. Mots indicateurs AVERTISSEMENT PRÉCAUTIONS Attention Remarque Pour une situation dangereuse avec un risque moyen pouvant être à l'origine d'un accident ou d'un décès, s'il n'est pas évité. Pour une situation dangereuse avec un faible risque pouvant être à l'origine de dommages au dispositif ou aux biens, d'une perte des données, ou d'un accident, si elle n'est pas évitée. Pour des informations importantes sur ce produit Pour des informations utiles sur le produit Symboles d'avertissement Danger général Risque d'explosion Tension alternative Danger de choc électrique Danger, produits corrosifs Courant continu FR-2 1.2 PH-mètre de laboratoire STARTER 5000 Précautions de sécurité ATTENTION Lire tous les avertissements de sécurité avant d’effectuer l’installation, les connexions ou les réparations de cet équipement. Si ces avertissements ne sont pas respectés, des accidents et/ou des dommages matériels peuvent en résulter. Conserver toutes les instructions pour référence ultérieure.           Vérifier que la plage de tension d'entrée de l'adaptateur CA et le type de prise sont compatibles avec l'alimentation alternative locale. S'assurer que le cordon d'alimentation n’est pas un obstacle potentiel ou ne présente aucun danger de chute. Utiliser l'équipement seulement dans des lieux secs. Sécher immédiatement tout déversement de liquide. L'instrument n'est pas étanche. Utiliser uniquement les accessoires et les périphériques approuvés. Faire fonctionner l'équipement uniquement dans les conditions ambiantes spécifiées dans ces instructions. Déconnecter l'équipement de l'alimentation électrique pour le nettoyer. Ne pas utiliser l'équipement dans des environnements dangereux ou instables. La maintenance doit être exclusivement exécutée par un personnel agréé. Si l'équipement est utilisé de façon non spécifiée par le fabricant, laprotection assurée par l'éq uipement peut être compromise. AVERTISSEMENT Ne pas utiliser cet instrument dans un environnement pouvant présenter des dangers d'explosion. L'enceinte de l'instrument n'est pas étanche aux gaz (danger d'explosion en présence d'étincelles, corrosion provoquée par l'introduction des gaz). ATTENTION Lors de l'utilisation de produits chimiques et de solvants, se conformer aux instructions du fabricant et aux règles générales de sécurité du laboratoire. PH-mètre de laboratoire STARTER 5000 2 FR-3 INSTALLATION Déballer soigneusement la boîte. 2.1 Contenu de l’emballage Le modèle ST5000-B (conditionnement de base) contient les articles suivants : ST5000-B pH-mètreST5000 Couvercle en utilisation ST5000 Porte-électrode autonome Clé USB 4Go Poudre pour tampon de pH (4,01 ; 7,00 ; 10,00) Alimentation Unités 1 1 1 ensemble 1 1 ensemble 1 ensemble En complément au contenu du ST500-B, l'emballage du ST5000-F contient aussi ce qui suit : Électrode de pH à remplissage en verre ST350 3 en 1 1 La poudre pour tampon de pH doit être dissoute dans 250 ml d'eau pure ou désionisée dans une fiole jaugée. Il est aussi possible de commander auprès d'Ohaus des solutions de tampons en bouteille. Contacter Ohaus pour de plus amples détails sur les électrodes, les solutions, les accessoires et les pièces de rechange. Modèle Description Référence ST350 Électrode de pH à remplissage en verre 3 en 1 30129354 ST310 Électrode de pH à remplissage en plastique 3 en 1 83033965 ST210 Électrode de pH à remplissage en plastique 2 en 1 83033966 Électrode de pH sous forme de gel en plastique 3 en 1 (nul 83033967 ST320 besoin de remplissage) ST260 Électrode de pH pour double pont salin en verre 2 en 1 30129357 ST230 Électrode de pH en verre pour échantillons troubles 2 en 1 83033968 STSURF Électrode de pH pour surface plane 2 en 1 30129470 STMICRO5 Électrode de pH pour microéchantillons 2 en 1 30087566 STMICRO8 Électrode de pH pour microéchantillons en verre 2 en 1 30087569 STTEMP30 Électrode de température 83033970 STORP1 Électrode ORP en plastique pour gel 30038555 STORP2 Électrode réutilisable ORP en verre 30038553 Poudre pour tampon (4,01 ; 7,00 ; 10,0) Tampon de pH 1,68 250 ml Tampon de pH 4,01 250ml Tampon de pH 7,00 250ml Tampon de pH 10,01 250ml Tampon de pH 12,45 250ml Électrolyte de référence pour électrode de pH Solution de protection pour électrode de pH Porte-électrode autonome Couvercle en utilisation ST5000 Adaptateur de câble SP RS232 F-M SF40A (pour ST5000) 83033971 30100424 30100425 30100427 30100429 30100440 30059255 30059256 30058733 30129897 30059316 FR-4 2.2 PH-mètre de laboratoire STARTER 5000 Installation du porte-électrode autonome Installer le bras de l'électrode sur la base : Après avoir réglé le bouton de tension jusqu'à un certain niveau, il est possible de déplacer le bras supérieur vers le haut et vers le bas. En plaçant l'électrode de pH sur le bras et en réalisant un étalonnage ou une mesure, il est possible de placer la bouteille de stockage de l'électrode de pH dans l'orifice de la base pour un rangement aisé: 2.3 Installation de l'adaptateur d'alimentation Insérer la prise de l'adaptateur dans la fente de l'adaptateur d'alimentation. ATTENTION N'utiliser qu'une prise d'adaptateur qui est compatible avec la prise secteur. S'assurer que la prise de l'adaptateur est bien enfoncée dans la fente de l'adaptateur d'alimentation avant de la connecter à la prise secteur. PH-mètre de laboratoire STARTER 5000 2.4 FR-5 Connexion des électrodes de pH Il existe 2 supports d'électrodes : support pH (BNC), support TEMP (Cinch). Pour l'électrode ST350 3 en 1, la connecter aux prises BNC et Cinch. Pour les électrodes de pH 2 en 1 (par exemple, ST230), les connecter seulement à la prise BNC. 3 ICÔNES ET AFFICHAGE 3.1 Mise sous tension de l’appareil Le STARTER 5000 est un pH-mètre de laboratoire 0,001 en couleurs et à écran tactile. Il ne comporte aucun bouton. Après l'avoir connecté à l'alimentation, le pH-mètre se mettra sous tension ; si la protection de l'écran est activée, appuyer sur l'écran pour l'activer. Remarque : Si le client n'utilise pas l'appareil pendant plus de 6 heures, ce dernier se mettra automatiquement hors tension. Il n'est plus possible d'appuyer sur l'écran pour le réactiver. Il doit être reconnecté à l'alimentation pour le remettre sous tension. Sélectionner la langue et appuyer ensuite sur Interface principale pour accéder à l'interface principale. FR-6 3.2 PH-mètre de laboratoire STARTER 5000 Mise hors tension de l'appareil Pour mettre l'appareil hors tension, appuyer sur le bouton de mise hors tension en partie supérieure gauche de l'interface principale pour accéder à l'interface de mise hors tension. Appuyer ensuite sur revenir à l'interface principale. pour mettre l'appareil hors tension, ou sur pour Mise hors tension de l'interface 3.3 Icônes ID utilisateur ID de l'échantillon Icône d'informations Date Heure Mode de mesure du pH. Passer entre les modes mV et Rel.mV ou Si l'icône de signal clignote, cela signifie que l'appareil reçoit un signal (de mesure ou d'étalonnage). Icônes de résultat final, de résultat final automatique, de résultat final manuel et de résultat final temporel. Avec un résultat final manuel, si le signal est stable, l'appareil affichera l'icône d'un signal stable. Appuyer sur l'icône d'arrêt Icône de signalisation stable Icône de température. pour atteindre le résultat final manuel. PH-mètre de laboratoire STARTER 5000 FR-7 Zone de touche des informations, des mesures et de l'étalonnage Cette zone d'informations sert aussi de raccourci pour entrer sur l'interface de paramétrage : Appuyer sur cette zone pour entrer sur l'interface afin de définir le mode de «Résultat final : résultat final. Les modes disponibles du résultat final sont Automatique, Manuel et Temporisé. Le mode de résultat final actuel est affiché. » Appuyer sur cette zone pour entrer sur l'interface et sélectionner ou modifier "Capteur ID :" l'ID du capteur. L'ID du capteur actuel est affichée. Toucher cette zone pour entrer sur l'interface des données d'étalonnage du "Condition capteur. L'icône des performances du capteur actuel (établies par rapport aux données d'étalonnage) est affichée. capteur :" Il existe 3 icônes de conditions du capteur : Pente : plus de 95 % et décalage : ± (0-15) mV Bonne condition de l'électrode “Tampons :” Pente : 90-95 % ou décalage : ± (15-35) mV Condition acceptable de l'électrode Pente : moins de 90 % ou décalage : ± (35-60) mV La condition de l'électrode n'est pas bonne ou nécessite un nettoyage Icône de rappel de l'étalonnage. Définir l’heure à laquelle l'appareil doit signaler que la durée depuis le dernier étalonnage est écoulée. Lorsque la durée est atteinte, cette icône s'affichera et clignotera pour rappeler qu'il est temps de réaliser un étalonnage. Cette zone affiche le groupe actuel de tampons, appuyer sur cette zone pour entrer sur l'interface de sélection du groupe de tampons. Icône de démarrage. Appuyer sur cette icône et l'appareil commencera la mesure, l'icône de signal clignotera. La zone de l'icône de démarrage affichera l'icône d'arrêt Icône d'étalonnage. Appuyer sur cette icône pour démarrer l'étalonnage. Icône d'enregistrement. Après avoir enregistré le résultat de la mesure en cours (résultat final atteint), la couleur de cette icône passera de noir à gris clair , ce qui indique que la mesure a déjà été enregistrée et ne peut l'être une deuxième fois. Icône de sortie. Diriger les données de sortie vers un dispositif externe tel qu'une imprimante. Si « Arrêt connexion » est choisi dans Paramétrage système ► Sortie, l'icône d'enregistrement même position. s'affichera. Sinon, l'icône de sortie s'affichera sur la FR-8 PH-mètre de laboratoire STARTER 5000 Retour à l'interface principale Retour au menu précédent Entrer, confirmer ou appliquer la configuration actuelle et terminer le paramétrage Modification Page précédente Page suivante Suppression Augmentation (éclairage) Diminution (éclairage) Affichage graphique (pour les données de mesurage ou d'étalonnage) Enregistrement Sortie vers un dispositif externe (clé USB, imprimante) Clavier complet Point d'étalonnage suivant Point d'étalonnage en cours Fin de l'étalonnage PH-mètre de laboratoire STARTER 5000 3.4 Répertoire du menu Le menu du premier et du second répertoire du STARTER 5000 : Date Heure Mode d'enregistrement Sortie Système Paramètre Luminosité de l'écran Débit en bauds Protection de l’écran Récupération des paramètres usine Informations sur l'instrument Mot passe maintenance Résolution Température MTC Saisie Mode pH/mV/Rel.mV Mesure Paramètre Mesure en continu Mode de résultat final Paramétrage ID Mode de mesure GLP Étalonnage Paramètre Données Groupe de tampons Rappel de l'étalonnage Données de mesure Données d'étalonnage FR-9 FR-10 4 PH-mètre de laboratoire STARTER 5000 ÉTALONNAGE, MESURE ET IMPRESSION L'affichage de l'appareil s'éteindra automatiquement conformément à la durée définie dans le menu Paramétrage système ►Protection de l'écran ► Arrêt automatique ACL. Pour remettre l'affichage sous tension, appuyer sur l'écran. (Voir 3.1) Le processus complet de mesure du pH doit suivre la procédure : a) Rincer l'électrode de pH et connecter à l'appareil b) Préparer le tampon et étalonner l'électrode de pH c) Préparer l'échantillon d) Réaliser la mesure du pH de l'échantillon e) Enregistrer les données de mesure f) Rincer l'électrode de pH et ranger Préparer l'électrode de pH : elle doit être rincée avec de l'eau pure avant et après utilisation. Vérifier si l'électrode est endommagée. (Prendre garde à l'ampoule en verre) L'électrode de pH doit être rangée dans une bouteille de stockage. Après avoir plongé l'électrode de pH dans l'échantillon ou dans la solution tampon, l'utilisateur doit homogénéiser pendant plusieurs secondes et attendre 30 à 60 secondes que le signal soit stable ; appuyer ensuite sur le bouton de fonctionnement (étalonnage ou mesure). AVERTISSEMENT Ne pas utiliser l'équipement dans des environnements dangereux. L'équipement n'est pas protégé contre les explosions. ATTENTION Avec des produits chimiques et des solvants, se conformer aux instructions du fabricant desdits produits et aux règles générales de sécurité du laboratoire. 4.1 Étalonnage 4.1.1 Groupe de tampons Le STARTER 5000 peut effectuer des étalonnages 1 et 5 points pour des groupes de tampons prédéfinis. Pour le groupe de tampons auto-définis, des étalonnages 1 à 9 points peuvent être réalisés. Avant l'étalonnage, s'assurer que le groupe actuel de tampons est bien celui désiré. Lors de l'utilisation d'une nouvelle électrode de pH, un étalonnage doit être réalisé avant d'effectuer des mesures. Si une électrode de pH reste inutilisée pendant quelques jours, un étalonnage doit être réalisé avant sa prochaine utilisation. L'étalonnage doit afficher la valeur correcte du pH lorsque l'appareil reçoit un signal en mV provenant de l'électrode de pH. Pente : le coefficient linéaire entre mV et pH conformément à la valeur théorique (par exemple, -59,16mV/pH à 25 °C signifie une pente de 100 %), Décalage : la valeur en mV lorsque la valeur du pH est égale à 7,00. (La valeur théorique est de 0 mV) PH-mètre de laboratoire STARTER 5000 FR-11 Le STARTER 5000 fournira une icône d'état de l'électrode afin de présenter les performances de l'électrode de pH par rapport aux données d'étalonnage. Normalement, si une icône à « visage triste » apparaît après un étalonnage adéquat, cela signifie qu'il faut remplacer l'électrode de pH par une nouvelle. Le STARTER 5000 corrige automatiquement la dépendance à la température des valeurs de pH du tampon. Par exemple, pour le groupe de tampons étalon US « 1,68, 4,01, 7,00, 10,01 », consulter le tableau suivant : 5 °C 1,67 4,01 7,09 10,25 10 °C 1,67 4,00 7,06 10,18 15 °C 1,67 4,00 7,04 10,12 20 °C 1,68 4,00 7,02 10,06 25 °C 1,68 4,01 7,00 10,01 30 °C 1,68 4,01 6,99 9,97 35 °C 1,69 4,02 6,98 9,93 40 °C 1,69 4,03 6,97 9,89 45 °C 1,70 4,05 6,97 9,86 50 °C 1,71 4,06 6,96 9,83 Par exemple, avec un tampon d'étalonnage 10,01 à 15 °C, le résultat de l'étalonnage doit être 10,12 pH. 4.1.2 Réalisation d'un étalonnage Après avoir rincé l'électrode de pH, plonger celle-ci dans le premier tampon, remuer pendant 10 secondes et appuyer ensuite sur pour démarrer l'étalonnage. Pendant l'étalonnage, l'icône du signal clignote. Lorsque le résultat final a été atteint, apparaît alors que disparaît, et le point d'étalonnage en cours est terminé. Il existe maintenant 3 options : Sortir de l'étalonnage et revenir à l'interface principale Terminer le point d'étalonnage en cours, se rendre sur l'interface des résultats de l'étalonnage Passer au point suivant d'étalonnage Remarque : Pendant le processus d'étalonnage, le ST5000 est toujours en Mode Résultat Final Automatique. Remarque : L'utilisation d'une électrode de pH 3 en 1 avec un capteur de température intégré ou une électrode de température séparée est recommandée. Avec une utilisation en mode MTC, entrer la valeur correcte de température et conserver tous les tampons et solutions d'échantillons à la température définie. FR-12 PH-mètre de laboratoire STARTER 5000 Mesure d'un échantillon 4.1.3 Mesure du pH S'assurer que l'électrode a été étalonnée avant la mesure. Après rinçage,  Placer l'électrode de pH dans l'échantillon.  Appuyer sur pour démarrer la mesure du pH, lorsque le résultat final a été atteint, la mesure en cours est terminée. Remarque : Si le signal est constamment instable, il est recommandé d'utiliser le mode de résultat final manuel afin d'éviter une mauvaise répétitivité. 4.1.4 Mesure en mV et Rel.mV Appuyer sur pour passer sur ou sur mode ; appliquer la même procédure que pour la mesure du pH afin de réaliser une mesure en mV ou en Rel.mV. Si l'appareil est connecté à une électrode ORP, le résultat en mV correspond à la valeur du potentiel d'oxydo-réduction, aussi connue sous le nom de valeur Redox. 4.2 Mesure de température Pour une précision supérieure, il est recommandé d'utiliser une électrode de température intégrée ou séparée.  Si une électrode de température est utilisée, ATC et la température de l'échantillon sont affichés.  Si l'appareil ne détecte pas d'électrode de température, il passe automatiquement au mode manuel de compensation de température et MTC apparaît. La température MTC doit être définie. PH-mètre de laboratoire STARTER 5000 4.3 FR-13 Impression Le STARTER 5000 peut imprimer la mesure ou les données d'étalonnage avec le port RS232. Après avoir correctement connecté l'imprimante (par exemple, OHAUS SF-40A) au pH-mètre, procéder aux paramétrages suivants (voir 5.1) : commander le câble de connexion du ST5000 (réf. 30059316) pour connecter les SF40A et ST5000. Paramétrage système ► Sortie, sélection Imprimante Paramétrage système ► Débit en bauds, sélectionner 9600. Le format de l'impression des mesures est le suivant : Date Heure Utilisateur ID échantillon 6-13-2015 14:20:22 Utilisateur01 Échantillon02 CapteurID Mode de résultat final MTC/ATC Capteur02 Manuel ATC Température Résultats 25,3 ºC 6,999 pH Le format de l'impression des données d'étalonnage est le suivant : Données d'étalonnage CapteurID Numéro de série Capteur02 14060012 Étalonnage Date Étalonnage Heure ATC/MTC 01-01-2015 14:20:33 ATC Tampons mV Décalage Pente 4,00 pH 176,94 mV 7,00 pH 0,05 mV 0,05 mV 98,22 % Lors de la réalisation d'une mesure en continu, le format des données d'étalonnage imprimées est le suivant : Synchronisation de l'enregistrement des mesures CapteurID Numéro de série Capteur03 14060012 Date Heure Température Mesure 01-01-2015 14:20:25 23,5 ºC 6,998 pH Date Heure Température Mesure 01-01-2015 14:20:30 23,5 ºC 6,999 pH Date Heure Température Mesure 01-01-2015 14:20:35 23,5 ºC 7,000 pH PH-mètre de laboratoire STARTER 5000 5 CONFIGURATION 5.1 Paramétrage système FR-14 Le paramétrage système est utilisé pour définir les paramètres de l'appareil, et les informations concernant ce dernier s'y trouvent aussi. Consulter la Section 3.4 pour des références. 5.1.1 Paramétrage système ► Mode Enregistrement Le mode Enregistrement possède 2 options, l'une concerne l'enregistrement automatique de la mesure du résultat final et l'autre concerne l'enregistrement manuel de la mesure du résultat final. « Enregistrement automatique de la lecture du résultat final » signifie que si la mesure atteint le résultat final (automatique, manuel ou temporisé), l'appareil enregistrera automatiquement la lecture du résultat final. L'icône d'enregistrement passera de noir à gris clair final ne peut être enregistrée qu'une seule fois. 5.1.2 . La mesure d'un résultat Paramétrage système ► Sortie Après avoir sélectionné l'arrêt de la connexion, l'interface principale affichera l'icône d'enregistrement 5.1.3 ou affichera l'icône de sortie Paramétrage système ► Restauration des paramètres par défaut La restauration selon la configuration usine comprend ce qui suit : 1. Mode GLP d'arrêt 2. Mode d'auto enregistrement avec arrêt automatique 3. Arrêt des mesures en continu 4. Arrêt de connexion (sortie) 5. Arrêt du rappel de l'étalonnage 6. La résolution est : 0,001 pH, 0,01mV 7. Mode de mesure : pH 8. Le groupe d'auto-tampons sera supprimé 9. Débit en bauds : 9600 10. Luminosité de l'écran : 70 % 11. Arrêt de la protection de l’écran 12. MTC 25 ºC 13. Résultat : Auto 14. Indicateur temporel du résultat final : 10 secondes 15. Supprimer toutes les données d'étalonnage 5.1.4 . Paramétrage système ► Mot de passe de service Non utilisé. PH-mètre de laboratoire STARTER 5000 5.2 Paramétrage des mesures 5.2.1 Paramétrage des mesures ► Mode de résultat final FR-15 Résultat final automatique Dans ce mode, l'appareil atteindra le résultat final et verrouillera automatiquement la mesure conformément aux critères suivants de stabilité : Résolution 0,1 pH : le signal de l'entrée du capteur peut ne pas être modifié de plus de 0,3 mV en 6 secondes. Résolution 0,01 pH : le signal de l'entrée du capteur peut ne pas être modifié de plus de 0,1 mV en 6 secondes. Résolution 0,001 pH : le signal de l'entrée du capteur peut ne pas être modifié de plus de 0,03 mV en 6 secondes. Résultat final temporel Dans ce mode, l'utilisateur doit introduire la durée du résultat final, par exemple, 15 secondes. Ceci signifie que 15 secondes après avoir débuté la mesure, l'appareil atteindra le résultat final. Résultat final Manuel Dans ce mode, l'utilisateur doit décider si la mesure du signal est stable, pour appuyer ensuite sur l'icône d'arrêt 5.2.2 et atteindre un résultat final manuel. Paramétrage des mesures ► Mode GLP de mesure Le mode GLP impose à chaque mesure de recevoir l'ID utilisateur et l'ID échantillon. En l'absence de l'un ou de l'autre, l'appareil affichera l'interface permettant à l'utilisateur d'introduire les données requises. 5.2.3 Paramétrage des mesures ► Mesure en continu Pendant une mesure en continu, l'utilisateur doit introduire l'intervalle temporel afin d'obtenir le résultat de chaque mesure. Le Graphique de la mesure en continu se trouve ci-dessous. L'axe des Y correspond à la valeur du pH. L'axe des X correspond au nombre de résultats de mesures. PH-mètre de laboratoire STARTER 5000 5.3 Paramétrage de l'étalonnage 5.3.1 Paramétrage de l'étalonnage ►Groupe de tampons FR-16 Avant l'étalonnage, s'assurer que le groupe actuel de tampons est correct pour les solutions de tampons. L'utilisateur peut sélectionner parmi 8 groupes de tampons prédéfinis. L'utilisateur peut aussi établir des groupes de tampons auto-définis. La différence entre 2 points d'étalonnage adjacent doit être supérieure à 0,99 pH. 5.3.2 Paramétrage de l'étalonnage ► Rappel de l'étalonnage Entrer ce réglage afin de définir la durée, depuis le dernier étalonnage, après combien d'heures l'appareil doit rappeler la conduite d'un nouvel étalonnage. L'appareil affichera l'icône de rappel de l'étalonnage 5.4 et clignotera afin de rappeler à l'utilisateur d'effectuer un étalonnage. Journal des données Appuyer sur Données pour entrer sur l'interface du journal des données, notamment celles des mesures et des étalonnages. Vous pouvez supprimer ici toutes les données de mesures. La restauration des paramètres par défaut reste le seul moyen de supprimer les données d'étalonnage. Les données peuvent être dirigées vers une clé USB (≤4 Go). PH-mètre de laboratoire STARTER 5000 6 MAINTENANCE 6.1 Message d’erreur FR-17 Lorsque la valeur introduite/le résultat de la mesure sont hors limites, ou si les données d'étalonnage ne sont pas suffisamment bonnes, l'appareil affichera des informations d'erreur. Si l'erreur ne peut pas être résolue en utilisant le message d'erreur affiché, contacter le représentant commercial d'Ohaus pour obtenir de l'aide. 6.2 Maintenance de l'appareil de mesure Les instruments de la série STARTER 5000 ne nécessitent aucune maintenance autre qu'un essuyage occasionnel avec un chiffon humide. Attention : Suivre ces précautions de sécurité lors du nettoyage de l'instrument.Débrancher l'instrument de sa prise d'alimentation.  Ne pas ouvrir l'instrument.  Ne laisser aucun liquide pénétrer dans l'instrument. Attention : L'enceinte est fabriquée en polystyrène-butadiène-acrylonitrile (ABS). Ce matériau est attaqué par certains solvants organiques comme le toluène, le xylène et l'éthyle méthyle cétone (MEK). Tout déversement doit être immédiatement asséché. 6.3 Maintenance de l'électrode S'assurer que l'électrode est remplie d'une solution électrolytique. Toujours stocker l'électrode conformément au manuel d'instructions de l'électrode ; ne pas la laisser s'assécher. Si la réponse de l'électrode devient lente ou si la pente n'est pas suffisamment bonne, essayer ce qui suit : o Tremper l'électrode dans de HCl 0,1M pendant plus de 8 heures. o Pour des contaminants huileux ou gras, dégraisser la membrane avec de la laine de coton trempée dans de l'acétone ou dans une solution savonneuse. Après avoir traité l'électrode, un nouvel étalonnage doit être réalisé. Si la pente de l'électrode n'est toujours pas bonne, l'électrode doit être remplacée. 6.4 Informations Appuyer sur l'icône Informations pour entrer sur l'interface d'informations. L'utilisateur peut ici introduire des informations pertinentes en appuyant sur la zone avec « » en bas et à gauche, telles que la date d'achat du capteur, etc. FR-18 PH-mètre de laboratoire STARTER 5000 7 DONNÉES TECHNIQUES 7.1 Spécifications Les données techniques sont valides dans les conditions ambiantes suivantes :  Utilisation en intérieur seulement  Altitude : Jusqu'à 2000 m  Plage de température spécifiée : 5 à 40°C  Humidité : humidité relative maximum à 80 % avec des températures jusqu'à 30°C diminuant linéairement à 50 % d'humidité relative à 40°C  Variations secteur jusqu'à ± 10 % de la tension nominale  Installation de catégorie II  Degré de pollution : 2  Fonctionnement assuré à température ambiante entre 5 et 40°C Modèle STARTER 5000 Plage de mesure -2,000…+ 20,000 pH –2000.00…+2000.00 mV -30 °C…+130 °C Résolution 0,1/0,01/0,001 pH 1/0,1/0,01 mV 0,1 °C Limites d'erreur ± 0,002 pH ± 0,03 % FS mV ± 0,1 °C Étalonnage 1-9 points 8 groupes de tampons prédéfinis + 1 groupe de tampons auto-définis 1 000 mesures Données d'étalonnage 10*10 Mémoire Alimentation Adaptateur CA, entrée 100-240 V~0,3 A input, sortie 9 V 1,12A Dimensions/poids Environ 210 x 170 x 80 mm/0,65 kg Poids/dimensions d'expédition Environ 372 x 310 x 185 mm/2,3 kg Affichage ACL tactile avec rétroéclairage Saisie BNC, impédance > > 3 x 10e+12 Ω Cinch, NTC 30 kΩ Compensation de température ATC et MTC Enceinte ABS PH-mètre de laboratoire STARTER 5000 8 FR-19 CONFORMITÉ Ce produit est conforme à la directive EMC 2004/108/EC et à la directive de basse tension 2006/95/EC. Sa Déclaration de conformité est disponible en ligne sur europe.ohaus.com/europe/en/home/support/compliance.aspx. Conformément à la directive européenne 2002/96/CE sur les déchets d’équipements électriques et électroniques (WEEE), cet appareil ne doit pas être mis au rebut comme des ordures ménagères. Ceci est également valable pour les pays en dehors de l’UE, selon leurs spécifications particulières. La directive 2006/66/EC sur les batteries a introduit de nouveaux impératifs depuis septembre 2008 sur le retrait des batteries d'un équipement destiné au rebut dans tous les États membres de l'Union européenne. Pour être conforme à cette directive, cet appareil a été conçu pour un retrait sans danger des batteries aussitôt sa fin de vie par une installation de traitement des déchets. Veuillez mettre au rebut ce produit conformément à la réglementation locale au point de collecte spécifié pour les équipements électriques et électroniques. En cas de questions, veuillez contacter l’autorité responsable ou le distributeur auprès duquel cet appareil a été acheté. Si cet appareil change de propriétaire (pour des raisons personnelles ou professionnelles), cette consigne doit lui être communiquée. Les instructions de mise au rebut en Europe sont disponibles en ligne sur europe.ohaus.com/europe/en/home/support/weee.aspx. Nous vous remercions de votre contribution à la protection de l’environnement. Remarque FCC : Cet équipement a été testé et déclaré conforme aux limites des appareils numériques de classe A, selon la Section 15 des règles du FCC (Commission fédérale des communications). Ces limites ont pour objectif de fournir une protection raisonnable contre des interférences dangereuses lorsque l'équipement est utilisé dans un environnement commercial. Cet équipement génère, utilise et peut rayonner des hautes fréquences et s'il n'est pas installé et utilisé conformément au guide de l'utilisateur, il peut générer des interférences préjudiciables aux communications radio. L'utilisation de cet équipement dans une zone résidentielle risque de générer des interférences préjudiciables, auquel cas l'utilisateur se verra dans l'obligation de rectifier la situation à ses frais. Remarque d’Industrie Canada Cet appareil numérique de classe A est conforme à la norme canadienne ICES-001. Enregistrement ISO 9001 En 1994, le Bureau Veritus Quality International (BVQI) a octroyé la certification d'enregistrement ISO 9001 à Ohaus Corporation, États-Unis d'Amérique, confirmant que le système de gestion de la qualité d'OHAUS était conforme aux spécifications normalisées de l'ISO 9001. Le 21 juin 2012, Ohaus Corporation, États-Unis d'Amérique, a été ré-enregistrée à la norme ISO 9001:2008. FR-20 9 PH-mètre de laboratoire STARTER 5000 GROUPES DE TAMPONS Le STARTER 5000 corrige automatiquement la dépendance àla température des valeurs de pH du groupe de tampons qui sont données dans les tableaux suivants. Le groupe de tampons doit être choisi avant l'étalonnage. JJG119 ((Réf. 25°C, 0,01pH) Temp °C 1,68 4,00 5 1,67 4,00 10 1,67 4,00 15 1,67 4,00 20 1,68 4,00 25 1,68 4,00 30 1,68 4,01 35 1,69 4,02 40 1,69 4,03 45 1,70 4,04 50 1,71 4,06 6,86 6,95 6,92 6,90 6,88 6,86 6,85 6,84 6,84 6,83 6,83 9,18 9,39 9,33 9,28 9,23 9,18 9,14 9,11 9,07 9,04 9,02 Normes des États-Unis (Réf, 25 °C) Temp °C 1,68 4,01 7,00 5 1,67 4,01 7,09 10 1,67 4,00 7,06 15 1,67 4,00 7,04 20 1,68 4,00 7,02 25 1,68 4,01 7,00 30 1,68 4,01 6,99 35 1,69 4,02 6,98 40 1,69 4,03 6,97 45 1,70 4,05 6,97 50 1,71 4,06 6,96 10,01 10,25 10,18 10,12 10,06 10,01 9,97 9,93 9,89 9,86 9,83 Norme européenne (Réf, 25 °C) Temp °C 2,00 4,01 7,00 5 2,02 4,01 7,09 10 2,01 4,00 7,06 15 2,00 4,00 7,04 20 2,00 4,00 7,02 25 2,00 4,01 7,00 30 1,99 4,01 6,99 35 1,99 4,02 6,98 40 1,98 4,03 6,97 45 1,98 4,04 6,97 50 1,98 4,06 6,97 9,21 9,45 9,38 9,32 9,26 9,21 9,16 9,11 9,06 9,03 8,99 12,46 13,21 13,01 12,82 12,64 12,46 12,29 12,13 11,98 11,83 11,70 11,00 11,72 11,54 11,36 11,18 11,00 10,82 10,64 10,46 10,28 10,10 PH-mètre de laboratoire STARTER 5000 MERCK (Réf, 20 °C) Temp °C 2,00 4,00 5 2,01 4,04 10 2,01 4,02 15 2,00 4,01 20 2,00 4,00 25 2,00 4,01 30 2,00 4,01 35 2,00 4,01 40 2,00 4,01 45 2,00 4,01 50 2,00 4,00 7,00 7,07 7,05 7,02 7,00 6,98 6,98 6,96 6,95 6,95 6,95 9,00 9,16 9,11 9,05 9,00 8,95 8,91 8,88 8,85 8,82 8,79 12,00 12,41 12,26 12,10 12,00 11,88 11,72 11,67 11,54 11,44 11,33 DIN (19267) (Réf, 25 °C) Temp °C 1,09 4,65 6,79 5 1,08 4,67 6,87 10 1,09 4,66 6,84 15 1,09 4,66 6,82 20 1,09 4,65 6,80 25 1,09 4,65 6,79 30 1,10 4,65 6,78 35 1,10 4,65 6,77 40 1,10 4,66 6,76 45 1,10 4,67 6,76 50 1,11 4,68 6,76 9,23 9,43 9,37 9,32 9,27 9,23 9,18 9,13 9,09 9,04 9,00 12,75 13,63 13,37 13,16 12,96 12,75 12,61 12,45 12,29 12,09 11,98 JIS Z 8802 (Réf, 25 °C) Temp °C 1,679 4,008 5 1,668 3,999 10 1,670 3,998 15 1,672 3,999 20 1,675 4,002 25 1,679 4,008 30 1,683 4,015 35 1,688 4,024 40 1,694 4,035 45 1,700 4,047 50 1,707 4,060 6,865 6,951 6,923 6,900 6,881 6,865 6,853 6,844 6,838 6,834 6,833 9,180 9,395 9,332 9,276 9,225 9,180 9,139 9,102 9,068 9,038 9,011 DIN(19266) / NIST (Réf, 25 °C) Temp °C 1,680 4,008 6,865 5 1,668 4,004 6,950 10 1,670 4,001 6,922 15 1,672 4,001 6,900 20 1,676 4,003 6,880 25 1,680 4,008 6,865 30 1,685 4,015 6,853 35 1,691 4,026 6,845 40 1,697 4,036 6,837 45 1,704 4,049 6,834 50 1,712 4,064 6,833 9,183 9,392 9,331 9,277 9,228 9,183 9,144 9,110 9,076 9,046 9,018 12,454 13,207 13,003 12,810 12,627 12,454 12,289 12,133 11,984 11,841 11,705 FR-21 FR-22 JJG119 (Réf, 25 °C, 0,001pH) Temp °C 1680 4,003 6,864 5 1,669 3,999 6,949 10 1,671 3,996 6,921 15 1,673 3,996 6,898 20 1,676 3,998 6,879 25 1,680 4,003 6,864 30 1,684 4,010 6,852 35 1,688 4,019 6,844 40 1,694 4,029 6,838 45 1,700 4,042 6,834 50 1,706 4,055 6,833 PH-mètre de laboratoire STARTER 5000 9,182 9,391 9,330 9,276 9,226 9,182 9,142 9,105 9,072 9,042 9,015 12460 13,210 13,011 12,820 12,637 12,460 12,292 12,130 11,975 11,828 11,697 PH-mètre de laboratoire STARTER 5000 FR-23 GARANTIE LIMITÉE Ohaus garantit que ses produits sont exempts de défauts matériels et de fabrication à compter de la date de livraison pendant toute la durée de la garantie, Pendant la période de garantie, Ohaus réparera ou, à sa convenance, remplacera sans frais toutes les pièces déterminées défectueuses, sous réserve que le produit soit retourné, fret payé d'avance, à Ohaus, Cette garantie ne s'applique pas si le produit a subi des dommages suite à un accident ou un mésusage, a été exposé à des matériaux radioactifs ou corrosifs, contient des matériaux étrangers ayant pénétré à l'intérieur ou après une maintenance ou des modifications apportées par quiconque autre qu’Ohaus, En l’absence d’une carte d'enregistrement de garantie dûment remplie, la période de garantie commence à la date de l'expédition au revendeur agréé, Aucune autre garantie expresse ou implicite n'est offerte par Ohaus Corporation, En aucun cas, Ohaus Corporation ne peut être tenu responsable des dommages indirects, Dans la mesure où les lois régissant les garanties varient d'un État à un autre et d'un pays à un autre, veuillez contacter Ohaus ou votre représentant local agréé Ohaus pour de plus amples informations, EN-22 STARTER 5000 Bench pH Meter Ohaus Corporation 7 Campus Drive Suite 310 Parsippany, NJ 07054 USA Tel: (973) 377-9000 Fax: (973) 944-7177 With offices worldwide/ Con oficinas en todo el mundo/ Avec des bureaux dans le monde entier www.ohaus.com *30129898* P/N 30129898 C© 2014 Ohaus Corporation, all rights reserved/ todos los derechos reservados/ tous droits réservés.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71