Wavetek 90 Manual de usuario

Tipo
Manual de usuario

Este manual también es adecuado para

TMD90TMD90
Dual Input
Digital
Thermometer
User Manual
Mode d'emploi
Bedienungshandbuch
Manuale d'Uso
Manual de uso
U.S. Service Center
Meterman Test Tools
1420 75th Street SW
Everett, WA 98203
Tel: 888-993-5853
Fax: 425-446-6390
Canadian Service Center
Meterman Test Tools
400 Britannia Rd. E. Unit #1
Mississauga, ON L4Z 1X9
Tel: 905-890-7600
Fax: 905-89-6866
European Correspondence Address*
Meterman Test Tools Europe
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
*Correspondence only - no repair or replacement
available from this address. European customers
please contact your distributor.
Visit www.metermantesttools.com for
• Catalog
• Application notes
• Product specifications
• Product manuals
PN 2099366
September 2003
© Wavetek Meterman Test Tools.
All rights reserved. Printed in China.
Please Recycle
®
®
®
®
1
TMD90 Dual Input Thermometer
Contents
Safety Information..............................................................................2
Symbols Used in this Manual...........................................................2
Introduction........................................................................................3
Display and Controls...........................................................................4
Functions............................................................................................ 5
Auto Power Off ( Sleep Mode )......................................................... 5
Change Thermocouple Type............................................................. 5
Using the Pushbuttons.....................................................................6
RS-232 Output................................................................................. 7
Troubleshooting.................................................................................. 8
Replacing the Battery.......................................................................... 9
9 V ac Adapter ....................................................................................9
Repair............................................................................................... 10
WARRANTY...................................................................................... 11
Thermocouple Definitions................................................................. 11
Specifications................................................................................... 12
Additional Specifications................................................................... 12
English
Document Dimensions: 3 x 5 inches
2
Safety Information
Place ONLY thermocouples in the thermocouple input.
Make sure your meter is configured for the thermocouple type to
be used.
Be sure the thermocouple you use can withstand the temperature
extreme it will be exposed to.
Properly maintain the meter and calibrate it regularly
Use the thermometer only as specified in this manual, or the
protection provided by the thermometer might be impaired.
Replace the battery as soon as BAT appears to avoid false readings
that can lead to electric shock and injury.
Warning
Supplied thermocouple is not intended for contact
with liquids or live electrical circuits.
Symbols Used in this Manual
Dangerous Voltage Refer to the manual
Double insulated Earth Ground
Complies with EU
directives
3
Introduction
The Model TMD90 dual input thermometer accepts Type K, J, T, R, S,
and E thermocouples. With a triple display screen to view measurement
results and relative time clock. Optional RS-232 software and cable
allows collection of real time data to a PC for further analysis or
reports.
TMD90 displays all LCD (liquid crystal display) segments for
approximately 3 seconds when it is first turned on and then provides
information on any connected thermocouples. Numerous viewing
combinations are available. The LCD is divided into three distinct
sections; one large (Primary) top screen and two smaller bottom
screens (Secondary and Relative Clock).
The three display areas are continually updated with the temperature
measurements and relative time information. There are several options
regarding how and what information is presented on the LCD.
Temperature readings are easily toggled between Fahrenheit and
Celsius.
A backlight illuminates the LCD for viewing in low light areas.
The button will freeze the upper display data while allowing
the lower displays to continue updating Information.
A low battery indicator is also displayed as appropriate.
The meter defaults to the last mode selected when turned off.
Note: If no thermocouples are connected, four dashes(----) appear in
the temperature display.
Box Contents
TMD90 Thermometer 1
Type K thermocouple probe 1
Users Manual 1
Hard plastic carry case 1
9 V battery (installed) 1
4
Display and Controls
10 (T2)
10 (T1)
5
4
32
6
7
1
8
9
11
6
12
13
REC Relative clock is active for recording min and max
temperature occurences or for time stamp purposes.
MAX/MIN Maximum or minimum temperature
measurements are being monitored.
HOLD/REL Freezes primary display or establishes a relative
zero for the primary display information.
BAT Low battery indicator.
°C °F Displays temperature in either degrees Celsius or
degrees Fahrenheit.
T1-T2 Toggles screen niformation from T1 (Primary)/ T 2
(Secondary) to T2 (Primary) / T1 (Secondary) ,then to T 1-T 2
(Differential) on Primary and alternating T1/T2 temperatures
on Secondary display.
Primary data display. Displays T1, T2, or T1-T2( temperature
differential -TD ) or a relative zero of T1, T2, or TD.
KJTRSE Selects proper input reference for thermocouple in
use. T1 and T2 must be the same thermocouple type.
Secondary data display. Displays T2,T1 or T1 and T2
temperatures.
Thermocouple input. Positive and negative polarized plugs for
the thermocouple probes used ( Blade type ); T1 on the left
,T2 on the right.
Relative clock display. Displays time in hours, minutes, and
seconds (HHMMSS) when REC is pressed and the relative
time that MIN or MAX data was recorded.
RS-232 output port. Optional software and cable – PN
TM-SW.
AC to DC converter input.
5
Caution
Read all Safety Information before using this
thermometer.
Functions
Auto Power Off (Sleep Mode)
The TMD90 shuts off automatically approximately 20 minutes after
being turned on. For recording or operating over longer periods of time
you can disable the sleep mode by pressing and
simultaneously while power on. When "n" then appear in the center of
the screen you can release the On. Auto power off is disabled when you
turn off the meter.
Change Thermocouple Type
Press and simultaneously for 2 or more seconds until K
appears. Press the button the type of probe cycles through K
(the default), J, T, R, S, and E types. The current mode is displayed on
the left side of the LCD. press , an "S" will appear in the center of
the screen.
6
Using the Pushbuttons
The display defaults to the mode last used. For your convenience the
meter defaults to the settings used during the last operation.
˚C ˚F
HOLD
REL
REC
T
1 T2
7
Use the pushbuttons to control operation of the TMD90. The table
assumes the TMD90 has been powered on with two thermocouples
installed and is set to display (default) T1 on the primary display, type
K thermocouple, and Record off.
Button Description
Turns the meter on and off.
Press and release HOLD/REL and the Primary display
(T1, T2, or T1-T2) freezes with HOLD displayed on top;
Press for two or more seconds REL appears on top of
LCD and the REL Primary display indicates the relative
zero. Relative zero causes the value of the primary
display to show as "000.0" , then only the amount of
temperature change will be indicated. Relative
temperatures can be recorded. Press HOLD/REL again
and the unit returns to default.
Press momentarily and the Primary display changes to
T2 (Secondary screen displays T1); press momentarily
again and it displays T1-T2; Secondary display
alternates between T1 and T2; Press momentarily again
and the instrument returns to default.
Press momentarily and the backlight illuminates for
approximately 30 seconds then turns off.
Press momentarily and the unit toggles between
Fahrenheit (the default) and Celsius temperatures; The
current mode is indicated on the right side of the LCD.
Press momentarily and the Relative Clock starts in the
lower right screen. All other button functions are locked
out except Power and Backlight. T1, T2, or T1-T2 is
displayed on the Primary screen; The Secondary screen
continues to update. Press momentarily again and the
unit cycles through MAX and MIN (Maximum and
Minimum recorded temperatures) and back to current
temperature; The record mode is displayed on the LCD.
Press and hold for three seconds to turn off the record
function.
RS-232 Output
With optional software and cable, P/N TM-SW, the TMD90 can output
measurement results to a computer with operating systems of Window
95/98/NT/2000/XP/ME. Features of the software:
Record up to 16,000 sample readings with real-time clock memory.
Programmable sample interval from 1 sec to 3,600 sec.
User defined maximum and minimum alarm settings.
Allow further analysis by downloading data to computer program.
All data is saved in .txt format for easy transformation into another
software program.
8
Troubleshooting
Problem Solution
Thermometer does not turn
on.
Check battery voltage and
replace discharged battery.
Verify that battery clips grip
the battery post tightly.
Dashes appear in the T1 and
T2 data screens.
Insert missing
thermocouple.
Dashes appear in T1 and/or T2
data screens with
thermocouples inserted.
Measure resistance of
thermocouples to ensure
they are not broken
internally.
Clean the thermocouple
and restart.
Temperature drifts from a
known value in a controlled
environment
Verify that thermocouple
type matches the displayed
icon.
Clean and dry the
thermocouple blades and
allow to air dry.
Confirm temperature with a
known good thermocouple.
Replace the defective
thermocouple if required.
9
Problem Solution
Relative clock does not start
when you push the
button.
Verify that the
thermocouple is inserted
correctly.
Dashes appear during a review
of the maximum recorded
value.
Thermocouple is not
acknowledged. Check for
intermittent or momentary
thermocouple removal.
Data continues to update after
you press the button.
Check for HOLD icon on
the display. Press
button firmly.
Instrument turns off while
recording.
Press and
simultaneously before
power on to disable the
Auto power off feature.
Replacing the Battery
Replace the 9 V battery when:
BAT appears on the right side of the display.
The meter will not turn on.
BAT appears when you turn on the backlight.
Even if the battery was recently replaced, check the voltage level if there
is no response from the meter.
To replace the battery
1. Remove all thermocouples from the top of the meter.
2. Lay the instrument face down on a clean, flat surface.
3. Loosen the screw and remove the battery door.
4. Replace the battery.
5. Replace the battery door and tighten the screw.
Remove the battery if you do not plan to use the meter for a month or
more. Do not leave the battery in a meter that may be exposed to
temperature extremes.
9 V ac Adapter
For long term measurement recording, a 9 V AC adapter can be used in
place of the battery. This can be purchased at any electronics store
using these specs: Miniature power plug with inner conductor positive
and outer conductor negative, 500 mA current rating. The 9 V battery
can be left in the TMD90 when using the AC adapter.
10
Repair
All test tools returned for warranty or non-warranty repair or for
calibration should be accompanied by the following: your name,
company’s name, address, telephone number, and proof of purchase.
Additionally, please include a brief description of the problem or the
service requested and include the test leads with the thermometer.
Non-warranty repair or replacement charges should be remitted in the
form of a check, a money order, credit card with expiration date, or a
purchase order made payable to Meterman Test Tools.
In-Warranty Repairs and Replacement – All Countries
Please read the warranty statement that follows, and check your
batteries and fuses before requesting repair. During the warranty
period any defective test tool can be returned to your Meterman Test
Tools distributor for an exchange for the same or like product. Please
check the “Where to Buy” section on www.metermantesttools.com for
a list of distributors near you. Additionally, in the United States and
Canada In-Warranty repair and replacement units can also be sent to a
Meterman Test Tools Service Center (see below for address).
Non-Warranty Repairs and Replacement – US and Canada
Non-warranty repairs in the United States and Canada should be sent to
a Meterman Test Tools Service Center. Call Meterman Test Tools or
inquire at your point of purchase for current repair and replacement
rates.
In USA In Canada
Meterman Test Tools Meterman Test Tools
1420 75th Street SW 400 Britannia Rd. E. Unit #1
Everett, WA 98203 Mississauga, ON L4Z 1X9
Tel: 888-993-5853 Tel: 905-890-7600
Fax: 425-446-6390 Fax: 905-890-6866
Non-Warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Meterman Test
Tools distributor for a nominal charge. Please check the “Where to
Buy” section on www.metermantesttools.com for a list of distributors
near you.
European Correspondence Address*
Meterman Test Tools Europe
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
*(Correspondence only – no repair or replacement available from this
address. European customers please contact your distributor.)
11
WARRANTY
The TMD90 Dual Input Thermometer is warranted against any defects
of material or workmanship within a period of one (1) year following
the date of purchase of the thermometer by the original purchaser or
original user. Any thermometer claimed to be defective during the
warranty period should be returned with proof of purchase to an
authorized Meterman Test Tools Service Center or to the local
Meterman Test Tools dealer or distributor where your thermometer
was purchased. See Repair section for details. Any implied warranties
arising out of the sale of a Meterman Test Tools thermometer,
including but not limited to implied warranties of merchantability and
fitness for a particular purpose, are limited in duration to the above
stated one (1) year period. Meterman Test Tools shall not be liable for
loss of use of the thermometer or other incidental or consequential
damages, expenses, or economical loss or for any claim or claims for
such damage, expenses or economical loss. Some states do not allow
limitations on how long implied warranties last or the exclusion or
limitation of incidental or consequential damages, so the above
limitations or exclusions may not apply to you. This warranty gives you
specific legal rights, and you may also have other rights which vary
from state to state.
Thermocouple Definitions
Type Alloy Alloy
J
Iron
Fe
Copper-Nickel
(Constantan)
Cu-Ni
K
Nickel-Chromium
(Chromel)
Ni-Cr
Nickel-Aluminum
Ni-AL
R
Platinum-13% Rhodium
Pt-13% Rh
Platinum
Pt
S
Platinum-10% Rhodium
Pt-10% Rh
Platinum
Pt
T
Copper
Cu
Copper-Nickel
(Constantan)
Cu-Ni
E
Nickel-Chromium
(Chromel)
Ni-Cr
Copper-Nickel
(Constantan)
Cu-Ni
12
Specifications
T/C Type Range Resolution Accuracy
K
-200 to 650 °C0.1 °C
651 to 1370 °C1.0 °C
(0.1% rdg + 0.7 °C)
-328 to 1000 °F0.1 °F
1001 to 2498 °F1.0 °F
(0.1% rdg + 1.4 °F)
J -200 to 500 °C0.1 °C
501 to 760 °C1.0 °C
(0.1% rdg + 0.7 °C)
-328 to 940 °F0.1 °C
941 to 1400 °F1.0 °F
(0.1% rdg + 1.4 °F
T
-200 to 390 °C0.1 °C (0.1% rdg + 0.7 °C)
-328 to 730 °F0.1 °F (0.1% rdg + 1.4 °F)
R/S
0 to 1000 °C0.1 °C
1001 to 1760 °C1.0 °C
(0.3% rdg + 0.7 °C)
32 to 1000 °F0.1 °F
1001 to 3200 °F1.0 °F
(0.1% rdg + 1.4 °F)
E
-200 to 380 °C0.1 °C
381 to 736 °C1.0 °C
(0.3% rdg + 0.7 °C)
-328 to 720 °F0.1 °F
721 to 1832 °F1.0°F
(0.1% rdg + 1.4 °F)
Additional Specifications
OPERATING CONDITIONS
0 to 50 °C (32 to 122 °F) at 0 to 85 % relative humidity
(non-condensing)
Ambient Coefficient
0 to 18 °C and 28 to 50 °C (Ambient temperatures) For each °C
ambient below 18 °C or above 28 °C, add the following tolerance into
the accuracy spec: 0.01% of reading +0.03 °C (0.01% of reading
+0.06 °F).
1
Thermomètre à double entrée TMD90
Table des matières
Consignes de sécurité..........................................................................2
Symboles utilisés dans ce mode d'emploi..........................................2
Introduction.........................................................................................3
Affichage et commandes......................................................................4
Fonctions.............................................................................................5
Arrêt automatique (mode de veille)....................................................5
Changer le type de thermocouple.......................................................5
Utilisation des boutons poussoirs......................................................6
Sortie RS-232...................................................................................8
Dépannage...........................................................................................9
Remplacement de la pile ....................................................................11
Adaptateur secteur de 9 V ..................................................................11
Réparation.........................................................................................12
GARANTIE.........................................................................................13
Définitions des thermocouples...........................................................13
Spécifications....................................................................................14
Autres caractéristiques.......................................................................14
Français
Français
2
Consignes de sécurité
Placer UNIQUEMENT des thermocouples dans l’entrée de
thermocouple.
S’assurer que l’appareil est configuré pour le type de thermocouple
concerné.
Vérifier que le thermocouple utilisé peut supporter les températures
extrêmes auxquelles il sera exposé.
Entretenir l’appareil avec soin et l’étalonner régulièrement
Utilisez uniquement le thermomètre, en respectant les indications
de ce manuel afin de ne pas entraver sa protection intégrée
Remplacer la pile dès l’apparition de l'indicateur de pile faible
apparaît pour éviter les mesures erronées qui posent des risques
d'électrocution et de blessures.
Avertissement
Le thermocouple fourni n’est pas conçu pour être en
contact avec des liquides ou des circuits électriques
sous tension.
Symboles utilisés dans ce mode d'emploi
Tension dangereuse
Se reporter au mode
d’emploi
Double isolation Prise de terre
Conforme aux
directives de l’UE
3
Introduction
Le thermomètre à entrée double modèle TMD90 accepte les
thermocouples de types K, J, T, R, S et E. Il est doté d’un écran à triple
affichage pour visualiser les résultats de mesure et l’horloge en temps
relatif. Le logiciel et le câble optionnel RS-232 permettent de recueillir
les données en temps réel vers un PC afin de les analyser ou de créer
ultérieurement des rapports.
Le TMD90 affiche tous les segments LCD (affichage à cristaux liquides)
pendant environ 3 secondes lors du démarrage avant de fournir les
informations sur tous les thermocouples connectés. Diverses
combinaisons d’affichage sont possibles. L’écran LCD est divisé en
trois sections distinctes ; une fenêtre principale (affichage primaire) en
haut et deux fenêtres plus petites en bas (affichage secondaire et
horloge relative).
Les trois zones d’affichage sont actualisées en permanence avec les
mesures de température et les informations de temps relatif. Plusieurs
options permettent de configurer la présentation et le type des
informations apparaissant sur l’écran LCD.
Les relevés de température basculent facilement entre les unités
Fahrenheit et Celsius.
Un rétroéclairage illumine l’écran LCD pour les lectures en zones
faiblement éclairées.
Le bouton gèle l’affichage de la fenêtre supérieure des
données tout en permettant la mise à jour des informations dans
les fenêtres du bas.
Un indicateur de pile faible apparaît également le cas échéant.
L’appareil adopte par défaut le dernier mode qui était sélectionné à
la mise hors tension.
Remarque : Si aucun thermocouple n’était connecté, quatre tirets (----)
apparaissent sur l’affichage des températures.
Contenu du coffret
Thermomètre TMD90 1
Sonde de thermocouple de type K 1
Mode d’emploi 1
Boîtier de transport en plastique rigide 1
Pile de 9 V (installée) 1
4
Affichage et commandes
10 (T2)
10 (T1)
5
4
32
6
7
1
8
9
11
6
12
13
REC L’horloge relative est active pour enregistrer les
incidences de température min et max ou pour
l’horodatage.
MAX/MIN Les mesures de température minimum ou
maximum actuellement contrôlées.
HOLD/REL Gèle l’affichage primaire ou établit un zéro
relatif pour les informations qu’elle contient.
BAT Témoin de pile faible.
°C °F Affiche la température en degrés Celsius ou
Fahrenheit.
T1-T2 Bascule entre l’affichage des informations T1
(primaire)/ T 2 (secondaire) et T2 (primaire) / T1
(secondaire), puis entre T 1-T 2(différentiel) sur l’affichage
principal et les températures alternées T1/T2 sur la fenêtre
secondaire.
Affichage des données primaires. Affiche T1, T2 ou T1-T2(
différentiel de température -TD ) ou un zéro relatif de T1, T2 ou TD.
KJTRSE Sélectionne la référence d’entrée appropriée pour le
thermocou ple utilisé. T1 et T2 doivent avoir le même type
de thermocouple.
Affichage des données secondaires. Affiche les
températures de T2,T1 ou de T1 et T2.
Entrée de thermocouple. Fiches à polarités positive et négative
pour les sondes de thermocouple utilisées (type de lame) ; T1 à
gauche ,T2 à droite.
Affichage d’horloge relative. Affiche l’heure en heures,
minutes et secondes (HHMMSS) à la pression de la touche
REC, et l’heure relative de l’enregistrement des résultats
MIN ou MAX.
5
Port de sortie RS-232. Câble et logiciel optionnel – Réf.TM-
SW.
Entrée du convertisseur de courant alternatif à courant
continu.
Attention
Lire toutes les consignes de sécurité avant d’utiliser
ce thermomètre.
Fonctions
Arrêt automatique (mode de veille)
Le TMD90 s’arrête automatiquement environ 20 minutes après sa mise
sous tension. Pour effectuer des enregistrements ou utiliser l’appareil
pendant de longues périodes, désactivez le mode de veille en appuyant
sur et sur simultanément avec l’appareil sous tension. A
l’apparition de « n » au centre de l’affichage, relâchez la touche On.
L’arrêt automatique est désactivé à la mise hors tension de l’appareil.
Changer le type de thermocouple
Appuyez sur le bouton et sur simultanément pendant au
moins 2 secondes jusqu’à l’apparition de K. Appuyez sur le bouton
et faites défiler les types de sondes, de K (par défaut) à J, T, R,
S, et E. Le mode actuel s’affiche dans le coin gauche de l’écran LCD.
Appuyez sur le bouton , un « S » apparaît au centre de l’écran.
6
Utilisation des boutons poussoirs
L’affichage adopte par défaut le dernier mode utilisé. Pour plus de
commodité, l’appareil utilise les paramètres utilisés pendant la dernière
opération.
˚C ˚F
HOLD
REL
REC
T
1 T2
Utilisez les boutons poussoirs pour contrôler le fonctionnement du
TMD 90. Dans le tableau suivant, on suppose que le TMD90 a été mis
sous tension avec deux thermocouples installés et qu’il est réglé pour
afficher (par défaut) T1 sur l’affichage primaire, le thermocouple de
type K et l’enregistrement inactif.
7
Bouton Description
Active ou désactive l’appareil.
Appuyez sur HOLD/REL, puis relâchez et
l'affichage principal (T1, T2 ou T1-T2) se fige
avec HOLD affiché en haut. Maintenez appuyer
pendant au moins deux secondes et REL
apparaît en haut de l'écran à cristaux liquides et
l'affichage principal REL indique le zéro relatif.
Le zéro relatif oblige l'affichage principal à
afficher la valeur « 000.0 », ensuite, seul le
changement de température sera indiqué. Les
températures relatives peuvent être enregistrées.
Appuyez à nouveau sur HOLD/REL et l'appareil
reprend sa valeur par défaut.
Une pression momentanée et l’affichage
primaire affiche T2 (la fenêtre secondaire affiche
T1) ; une nouvelle pression momentanée permet
d’afficher T1-T2 ; la fenêtre secondaire bascule
entre T1 et T2 ; une dernière pression
momentanée et l’instrument revient au réglage
par défaut.
Une pression momentanée et le rétroéclairage
s’allume pendant une trentaine de secondes
avant de s’éteindre.
Une pression momentanée et l’appareil bascule
entre les températures en Fahrenheit (par
défaut) et Celsius ; le mode courant est indiqué
à droite sur l’écran LCD.
Une pression momentanée et l’horloge relative
(REC) démarre dans la fenêtre inférieure droite,
s’affiche dans le coin supérieur gauche de
l’écran LCD – Toutes les autres fonctions sont
verrouillées en dehors de l’alimentation et du
rétroéclairage. T1, T2 ou T1-T2 s’affiche sur
l’affichage primaire ; la fenêtre secondaire
continue d’être actualisée. Une nouvelle
pression momentanée et l’appareil répète les
valeurs MAX et MIN (températures maximum et
minimum enregistrées) et revient à la
température courante ; le mode
d’enregistrement apparaît sur l’écran LCD.
Maintenez la touche enfoncée pendant trois
secondes pour désactiver la fonction
d’enregistrement.
8
Sortie RS-232
L’utilisation d’un câble et d’un logiciel optionnel, réf. T90/10-SW,
permet au TMD90 de transmettre les résultats des mesures vers un
ordinateur fonctionnant sous Windows 95/98/NT/2000/XP/ME.
Fonctionnalités du logiciel :
Enregistre jusqu’à 16.000 valeurs exemples en mémoire avec horloge en
temps réel.
Intervalle d’échantillonnage programmable de 1 s à 3600 s.
Paramètres d’alarme maximum et minimum définis par l’utilisateur.
Permet une analyse approfondie en téléchargeant les données vers
un programme informatique.
Toutes les données sont enregistrées en format .txt permettant de
les convertir facilement dans un autre programme d’application.
9
Dépannage
Problème Solution
Le thermomètre ne s’active
pas.
Vérifiez la tension de la pile
et remplacez la pile si elle
est déchargée.
Assurez-vous que les
pinces enserrent
étroitement la borne de la
pile.
Des tirets apparaissent dans le
fenêtres de résultats T1 et T2.
Insérez le thermocouple
manquant.
Des tirets apparaissent dans
les fenêtres de résultats T1
et/ou T2 avec les
thermocouples insérés.
Mesurez la résistance des
thermocouples pour vérifier
s’ils ne présentent pas de
cassure interne.
Nettoyez le thermocouple
et recommencez.
10
Problème Solution
La température dérive d’une
valeur connue en
environnement contrôlé
Vérifiez si le type de
thermocouple correspond
à l’icône affichée.
Nettoyez et séchez les
lames de thermocouple et
laissez sécher à l’air libre.
Confirmez la température
avec un thermocouple
identifié correct.
Remplacez le
thermocouple défectueux
s’il y lieu.
L’horloge relative ne démarre
pas à la pression du bouton
.
Vérifiez si le thermocouple
est inséré correctement.
Des tirets apparaissent
pendant l’examen de la valeur
maximale enregistrée.
Le thermocouple n'est pas
reconnu. Vérifiez l’absence
de retraits intermittents ou
momentanés du
thermocouple.
Les données continuent d’être
actualisées après la pression
du bouton .
Recherchez l'icône HOLD
sur l'affichage. Appuyez
fermement sur le bouton
.
L’appareil s’éteint pendant
l’enregistrement.
Appuyez sur et sur
simultanément avant
le démarrage pour
désactiver la fonction de
mise en veille automatique.
11
Remplacement de la pile
Remplacez la pile de 9 V lorsque :
BAT apparaît à droite sur l’affichage.
L’appareil ne s’allume pas.
BAT apparaît au démarrage du rétroéclairage.
Même si la pile vient d’être remplacée, vérifiez son niveau de tension si
l’appareil ne réagit pas.
Pour remplacer la pile
1. Retirez tous les thermocouples du sommet de l’appareil.
2. Posez l’appareil face vers le bas sur une surface propre et plane.
3. Desserrez la vis et enlevez le capot de pile.
4. Remplacez la pile.
5. Replacez le capot de pile et serrez la vis.
Retirez la pile si vous n’envisagez pas d’utiliser l’appareil pendant plus
d’un mois. Ne laissez pas la pile dans l’appareil s’il risque d’être exposé
à des températures extrêmes.
Adaptateur secteur de 9 V
Un adaptateur secteur de 9 V peut être utilisé à la place de la pile pour
enregistrer les mesures à long terme. Cet adaptateur vendu dans les
magasins d’électronique a les caractéristiques suivantes : un
connecteur d’alimentation miniature doté d’un conducteur interne
positif et d’un conducteur externe négatif, une intensité de 500mA. La
pile de 9V peut être laissée dans le TMD90 pendant l’utilisation de
l’adaptateur secteur.
12
Réparation
Tous les outils de test renvoyés pour un étalonnage ou une réparation
couverte ou non par la garantie doivent être accompagnés des éléments
suivants : nom, raison sociale, adresse, numéro de téléphone et justificatif
d'achat. Ajoutez également une brève description du problème ou du service
demandé et incluez les cordons de test avec le thermomètre. Les frais de
remplacement ou de réparation hors garantie doivent être acquittés par
chèque, mandat, carte de crédit avec date d'expiration, ou par bon de
commande payable à l'ordre de Meterman Test Tools.
Remplacements et réparations sous garantie – Tous pays
Veuillez lire la déclaration de garantie qui suit, et vérifiez les piles, batteries et
fusibles avant de demander une réparation. Pendant la période de garantie,
tout outil de test défectueux peut être renvoyé auprès de votre distributeur
Meterman Test Tools pour être échangé contre un produit identique ou
similaire. Consultez la section « Where to Buy » sur le site
www.metermantesttools.com pour obtenir la liste des distributeurs dans
votre région. Au Canada et aux Etats-Unis, les appareils devant être remplacé
ou réparé sous garantie peuvent également être envoyés dans un centre de
services (voir les adresses ci-dessous).
Remplacements et réparations hors garantie – Canada et
Etats-Unis
Les appareils à réparer hors garantie au Canada et aux Etats-Unis doivent être
envoyés dans un centre de services Meterman Test Tools. Appelez Meterman
Test Tools ou renseignez-vous auprès de votre lieu d’achat pour connaître les
tarifs en vigueur de remplacement ou de réparation.
Aux Etats-Unis Au Canada
Meterman Test Tools Meterman Test Tools
1420 75th Street SW 400 Britannia Rd. E. Unit #1
Everett, WA 98203 Mississauga, Ontario L4Z 1X9
Tél. : 888-993-5853 Tél. : 905-890-7600
Fax : 425-446-6390 Fax : 905-890-6866
Remplacements et réparations hors garantie – Europe
Les appareils européens non couverts par la garantie peuvent être remplacés
par votre distributeur Meterman Test Tools pour une somme nominale.
Consultez la section « Where to Buy » sur le site
www.metermantesttools.com pour obtenir la liste des distributeurs dans
votre région.
Adresse postale européenne*
Meterman Test Tools Europe
P.O. Box 1186
5602 B.D. Eindhoven
Pays-Bas
*(Réservée à la correspondance – Aucune réparation ou remplacement n’est
possible à cette adresse. Nos clients européens doivent contacter leur
distributeur.)
13
GARANTIE
Le thermomètre à double entrée TMD90 est garanti contre tout défaut
de fabrication ou de main d’œuvre pendant une période d’un (1) an à
compter de la date d’achat du thermomètre par l’acheteur initial ou
l’utilisateur initial. Tout thermomètre faisant l'objet d'un défaut pendant
la période de garantie doit être renvoyé accompagné d’un justificatif
d’achat auprès d’un centre de services agréé par Meterman Test ou du
distributeur ou du revendeur local de Meterman Test Tools où l'achat
du thermomètre a été effectué. Voir la section Réparation pour tous les
détails. Toutes les garanties implicites résultant de la vente d’un
thermomètre Meterman Test Tools, y compris mais sans s’y limiter les
garanties de commercialisation ou d’adaptation à un usage particulier,
sont limitées à la durée d’un (1) an déjà citée. Meterman Test Tools ne
sera pas tenu responsable de la privation de jouissance du
thermomètre ou d’autres dommages directs ou indirects, frais ou
pertes économiques ni des poursuites engagées pour de tels
dommages, frais ou pertes économiques. Certains pays n’admettent
pas les limitations sur la durée des garanties implicites, ni sur
l’exclusion ou la limitation des dommages directs ou indirects ; il est
donc possible que les limitations ou exclusions de cette garantie ne
s’appliquent pas dans votre cas. La présente garantie confère certains
droits juridiques : la législation du pays ou de l’état peut vous en
accorder d’autres.
Définitions des thermocouples
Type Alliage Alliage
J
Fer
Fe
Cupro-Nickel
(Constantan)
Cu-Ni
K
Nickel-Chrome
(Chromel)
Ni-Cr
Nickel-Aluminium
Ni-AL
R
Platinum-13 %
Rhodium
Pt-13% Rh
Platinum
Pt
B
Platinum-10 %
Rhodium
Pt-10% Rh
Platinum
Pt
T
Cuivre
Cu
Cupro-Nickel
(Constantan)
Cu-Ni
E
Nickel-Chrome
(Chromel)
Ni-Cr
Cupro-Nickel
(Constantan)
Cu-Ni
14
Spécifications
Type T/C Gamme Résolution Précision
K
-200 à 650 °C0.1 °C
651 à 1370 °C1.0 °C
(0,1
% de lecture + 0,7
°C
)
-328 à + 1000 °F0,1 °F
1001 à 2498 °F1,0 °F
(0,1 % de lecture +1,4 °F)
J
-200 à 500 °C0,1 °C
501 à 760 °C1,0 °C
(0,1
% de lecture + 0,7
°C
)
-328 à 940 °F0,1 °C
941 à 1400 °F1,0 °F
(0,1 % de lecture +1,4 °F)
T -200 à 390 °C0,1 °C
(0,1
% de lecture + 0,7
°C
)
-328 à 730 °F0,1 °F
(0,1 % de lecture +1,4 °F)
R/S 0 à 1000 °C0,1 °C
1001 à 1760 °C1,0 °C
(0,3% de lecture + 0,7 °C)
32 à 1000 °F0.1 °F
1001 à 3200 °F1.0 °F
(0,1 % de lecture +1,4 °F)
E
-200 à 380 °C0.1 °C
381 à 736 °C1.0 °C
(0,3% de lecture + 0,7 °C)
-328 à 720 °F0.1 °F
721 à 1832 °F1.0°F
(0,1 % de lecture +1,4 °F)
Autres caractéristiques
CONDITIONS D’EXPLOITATION
0 à 50 °C (32 à 122 °F) avec 0 à 85 % d’humidité relative (sans
condensation)
Coefficient ambiant
de 0 à 18 °C et de 28 à 50 °C (températures ambiantes) Pour chaque
°C ambiant inférieur à 18 °C ou supérieur à 28 °C, ajouter la tolérance
suivante à la caractéristique de précision : 0,01 % de lecture + 0,03 °C
(0,01 % de lecture + 0,06 °F)
1
TMD90 Dual Input Thermometer
Inhalt
Sicherheitsinformationen .....................................................................2
Symbole in diesem Handbuch...........................................................2
Einleitung.............................................................................................3
Anzeige und Bedienelemente................................................................4
Funktionen...........................................................................................5
Automatische Abschaltung (Ruhemodus)..........................................5
Ändern des Thermoelementtyps........................................................5
Verwenden der Tasten.......................................................................6
RS-232-Ausgang ..............................................................................8
Fehlerbehebung...................................................................................9
Ersetzen der Batterie..........................................................................11
9-V-Wechselstromadapter..................................................................11
Reparatur ..........................................................................................12
GARANTIE.........................................................................................13
Thermoelementdefinitionen................................................................13
Spezifikationen...................................................................................14
Weitere Spezifikationen......................................................................14
Deutsch
Deutsch
2
Sicherheitsinformationen
NUR Thermoelemente in den Thermoelementeingang einführen.
Sicherstellen, dass das Messgerät für den verwendeten
Thermoelementtyp konfiguriert ist.
Prüfen, ob das verwendete Thermoelement die Temperaturextreme,
denen es ausgesetzt wird, aushalten kann.
Das Messgerät ordnungsgemäß unterhalten und regelmäßig
kalibrieren.
Das Thermometer ausschließlich wie in diesem Handbuch
beschrieben einsetzen, da sonst die im Thermometer integrierten
Schutzeinrichtungen beeinträchtigt werden können.
Zur Vermeidung falscher Messwerte, die zu Stromschlag oder
Verletzungen führen können, die Batterie ersetzen, sobald die BAT
eingeblendet wird.
Warnung
Das gelieferte Thermoelement darf nicht mit
Flüssigkeiten oder stromführenden Stromkreisen in
Kontakt kommen.
Symbole in diesem Handbuch
Gefährliche Spannung Im Handbuch nachlesen.
Schutzisoliert Erde, Masse
Übereinstimmung mit
EU-Richtlinien
3
Einleitung
Das Modell TMD90 Dual Input Thermometer akzeptiert die
Thermoelementtypen K, J, T, R, S und E. Das Gerät verfügt über einen
Dreifach-Anzeigebildschirm zur Anzeige der Messergebnisse und des
relativen Zeitgebers. Wahlfreie RS-232-Software und Kabel
ermöglichen das Aufzeichnen von Echtzeitdaten auf einem PC für
spätere Analysen und Berichte.
Das TMD90 zeigt nach dem Einschalten alle LCD-Segmente (LCD =
liquid crystal display) ungefähr 3 Sekunden lang an, und liefert dann
Informationen zu beliebigen angeschlossenen Thermoelementen. Es
gibt mehrere Anzeigekombinationen. Die LCD ist in drei gesonderte
Abschnitte unterteilt: eine große obere Anzeige (Primäranzeige) und
zwei kleinere untere Anzeigen (Sekundäranzeige und Anzeige für den
relativen Zeitgeber).
Die drei Anzeigebereiche werden kontinuierlich mit Informationen zu
Temperaturmessungen und zum relativen Zeitgeber aktualisiert. Es gibt
mehrere Optionen bezüglich Anzeige und Auswahl der auf der LCD
dargestellten Informationen.
Temperaturmesswerte können einfach zwischen Fahrenheit und
Celsius umgeschaltet werden.
Eine Hintergrundbeleuchtung beleuchtet die LCD für Ablesungen in
Arbeitsbereichen mit wenig Licht.
Die Taste friert die oberen Anzeigedaten ein und ermöglicht
fortgesetzte Aktualisierung der Informationen der unteren Anzeigen.
Darüber hinaus wird ggf. auch eine Anzeige für schwache Batterie
angezeigt.
Das Messgerät kehrt beim Einschalten zu dem Modus zurück, der
vor dem Ausschalten zuletzt verwendet wurde.
Hinweis: Wenn keine Thermoelemente angeschlossen sind, werden in
der Temperaturanzeige vier Gedankenstriche (----) angezeigt.
Lieferumfang
TMD90 Thermometer 1
Typ K Thermoelementsonde 1
Bedienungsanleitung 1
Transportbehälter 1
9 V Batterie (installiert) 1
4
Anzeige und Bedienelemente
10 (T2)
10 (T1)
5
4
32
6
7
1
8
9
11
6
12
13
REC Relativer Zeitgeber aktiviert zur Aufzeichnung der
Min.- und Max.-Temperatur und für Zeitstempelvergabe.
MAX/MIN Min.- bzw. Max.-Temperaturmesswerte werden
überwacht.
HOLD/REL Friert Primäranzeige ein bzw. setzt einen
relativen Nullpunkt für die Daten in der Primäranzeige.
BAT Anzeige für schwache Batterie.
°C °F Zeigt Temperatur in Grad Celsius oder Grad
Fahrenheit an.
T1-T2 Schaltet Anzeigeinformationen um: von T1
(Primäranzeige)/ T2 (Sekundäranzeige) zu T2
(Primäranzeige) / T1 (Sekundäranzeige), dann zu T1-T2
(Differenzial) auf der Primäranzeige und wechselweise
T1/T2-Temperaturen auf der Sekundäranzeige.
Primärdatenanzeige. Zeigt T1, T2 oder T1-T2
(Temperaturdifferenzial -TD ) oder einen relativen
Nullpunkt von T1, T2 oder TD an.
KJTRSE Wählt korrekte Eingangsreferenz für das
verwendete Thermoelement aus. T1 und T2 müssen
denselben Thermoelementtyp aufweisen.
Sekundärdatenanzeige. Zeigt T2, T1 oder T1 und T2
Temperaturen an.
Thermoelementeingang. Positiv und negativ polarisierte
Stecker für die verwendeten Thermoelementsonden
(Spatenstecker); T1 links, T2 rechts.
Anzeige des relativen Zeitgebers. Zeigt die Uhrzeit in
Stunden, Minuten und Sekunden (HHMMSS) an, wenn
REC gedrückt wird, und die relative Zeit der Aufzeichnung
der MIN- bzw. MAX-Daten an.
5
RS-232-Ausgangsanschluss.
Wahlfreie Software und Kabel – Teile-Nr. TM-SW.
Wechselstrom-/Gleichstromwandlereingang.
Vorsicht
Vor Gebrauch dieses Thermometers alle
Sicherheitsinformationen lesen.
Funktionen
Automatische Abschaltung (Ruhemodus)
Das TMD90 schaltet ungefähr 20 Minuten nach dem Einschalten
automatisch ab. Für Aufzeichnungen bzw. Betrieb über längere
Zeitperioden kann der Ruhemodus durch gleichzeitiges Drücken der
Tasten und während des Einschaltens deaktiviert werden.
Wenn „n“ in der Mitte der Anzeige erscheint, kann die ON-Taste
losgelassen werden. Die automatische Abschaltung wird deaktiviert,
wenn das Messgerät ausgeschaltet wird.
Ändern des Thermoelementtyps
Die Tasten und gleichzeitig mindestens 2 Sekunden lang
drücken, bis K angezeigt wird. Die Taste drücken, sodass die
möglichen Sondentypen durchlaufen werden: K (Standard), J, T, R, S
und E. Der aktuelle Modus wird auf der linken Seite der LCD angezeigt.
drücken, sodass ein „S“ in der Mitte der Anzeige erscheint.
6
Verwenden der Tasten
Die Anzeige kehrt standardmäßig zum zuletzt verwendeten Modus
zurück. Das Messgerät verwendet standardmäßig die Einstellungen, die
während des letzten Einsatzes verwendet wurden.
˚C ˚F
HOLD
REL
REC
T
1 T2
7
Die Tasten zur Steuerung des TMD90 verwenden. Die Tabelle setzt
voraus, dass das TMD90 über zwei installierte Thermoelemente
verfügt, eingeschaltet ist, zur Anzeige von T1 (Standard) auf der
Primäranzeige und Typ K Thermoelement eingestellt ist und „Record“
deaktiviert ist.
Taste Beschreibung
Schaltet das Messgerät ein und aus.
Kurzzeitiges Drücken von HOLD/REL bewirkt, dass die
Primäranzeige (T1, T2 bzw. T1-T2) einfriert, und HOLD
wird oben angezeigt. Drücken für 2 oder mehr Sekunden
bewirkt, dass REL oben in der LCD angezeigt wird, und die
REL-Primäranzeige zeigt den relativen Nullpunkt an.
Relativer Nullpunkt bewirkt, dass der Wert der
Primäranzeige als "000,0" angezeigt wird, danach wird nur
die Temperaturveränderung angezeigt. Relative
Temperaturen können aufgezeichnet werden.
Nochmaliges Drücken von HOLD/REL bewirkt, dass die
Einheit auf Standard zurückschaltet.
Kurzzeitiges Drücken bewirkt, dass die
Primäranzeige auf T2 (Sekundäranzeige zeigt T1 an)
schaltet; nochmaliges kurzzeitiges Drücken bewirkt,
dass T1-T2 angezeigt wird; Sekundäranzeige
wechselt zwischen T1 und T2; nochmaliges
kurzzeitiges Drücken bewirkt, dass das Instrument
auf Standard zurückschaltet.
Kurzzeitiges Drücken bewirkt, dass die
Hintergrundbeleuchtung ungefähr 30 Sekunden lang
leuchtet und dann ausschaltet.
Kurzzeitiges Drücken bewirkt, dass die Einheit
zwischen Fahrenheit- (Standard) und Celsius-
Temperaturen umschaltet; der aktuelle Modus
rechts auf der LCD angezeigt.
Kurzzeitiges Drücken bewirkt, dass der relative
Zeitgeber unten rechts auf der Anzeige zu laufen
beginnt; ... wird oben links auf der LCD angezeigt -
alle übrigen Tastenfunktionen mit Ausnahme von
„Power“ und „Backlight“ sind ausgesperrt. T1, T2
bzw. T1-T2 wird in der Primäranzeige angezeigt; die
Sekundäranzeige wird weiterhin aktualisiert.
Nochmaliges kurzzeitiges Drücken bewirkt, dass die
Einheit die MAX- und MIN-Werte (aufgezeichnete
Maximal- und Minimaltemperaturen) durchläuft und
zur aktuellen Temperaturanzeige zurückkehrt. Die
Taste drücken und 3 Sekunden halten, um die
Aufzeichnungsfunktion auszuschalten.
8
RS-232-Ausgang
Mit wahlfreier Software und Kabel, Teile-Nr. T90/10-SW, kann das
TMD90 Messergebnisse auf einen Computer unter Windows
95/98/NT/2000/XP/ME ausgeben. Zu den Merkmalen der Software
gehören:
Aufzeichnung von bis zu 16000 Messwerten in Speicher mit
Echtzeituhr.
Programmierbares Intervall: 1 bis 3600 Sekunden.
Benutzerdefinierte Maximal- und Minimal-Alarmeinstellungen.
Ermöglicht weitere Analysen durch Herunterladen von Daten auf
einen Computer Verarbeitung mit einem Programm.
Alle Daten werden im .txt-Format gespeichert, sodass sie einfach in
andere Softwareprogramme übernommen werden können.
9
Fehlerbehebung
Problem Lösung
Thermometer lässt sich nicht
einschalten.
Batteriespannung prüfen
und ggf. aufgebrauchte
Batterie ersetzen.
Sicherstellen, dass
Batterieklemmen die
Batteriepole sicher
festhalten.
Gedankenstriche erscheinen in
den T1- und T2-
Datenanzeigen.
Fehlendes Thermoelement
einsetzen.
Gedankenstriche erscheinen in
den T1- und/oder T2-
Datenanzeigen und
Thermoelemente sind
eingesetzt.
Widerstand der
Thermoelemente messen,
um interne Defekte
auszuschließen.
Das Thermoelement
reinigen und neu beginnen.
10
Problem Lösung
Temperatur schwankt in einer
kontrollierten Umgebung von
einem bekannten Wert.
Prüfen, ob
Thermoelementtyp mit dem
angezeigten Symbol
übereinstimmt.
Die Thermoelementblätter
reinigen und trocknen und
die Einheit lufttrocknen
lassen.
Temperatur mit einem
einwandfreien
Thermoelement bestätigen.
Das defekte
Thermoelement
nötigenfalls ersetzen.
Der relative Zeitgeber startet
nicht, wenn die Taste
gedrückt wird.
Sicherstellen, dass das
Thermoelement korrekt
eingesetzt ist.
Gedankenstriche erscheinen
während der Anzeige des
aufgezeichneten
Maximalwerts.
Thermoelement ist nicht
bestätigt. Auf
zeitweilige/kurzzeitige
Trennung des
Thermoelements prüfen.
Daten werden nach Drücken
der Taste weiterhin
aktualisiert.
Prüfen, ob das HOLD-
Symbol HOLD auf der
Anzeige erscheint. Die
Taste fest drücken.
Instrument schaltet sich
während einer Aufzeichnung
ab.
Die Tasten und
vor dem Einschalten
gleichzeitig drücken, um
die Funktion „Automatische
Abschaltung“ zu
deaktivieren.
11
Ersetzen der Batterie
Die 9-V-Batterie ersetzen, wenn:
BAT rechts auf der Anzeige eingeblendet wird.
Das Messgerät nicht eingeschaltet werden kann.
BAT eingeblendet wird, wenn die Hintergrundbeleuchtung aktiviert
wird.
Selbst wenn die Batterie kürzlich ersetzt wurde, den Spannungspegel
prüfen, falls das Messgerät nicht reagiert.
Ersetzen der Batterien
1. Alle Thermoelemente vom Oberteil des Messgeräts entfernen.
2. Das Instrument mit der Vorderseite nach unten auf eine saubere,
flache Oberfläche legen.
3. Die Schraube lösen und die Batteriefachabdeckung abnehmen.
4. Die Batterie ersetzen.
5. Die Batteriefachabdeckung wieder anbringen und die Schraube
anziehen.
Die Batterie entfernen, falls das Messgerät voraussichtlich einen Monat
oder länger nicht verwendet wird. Die Batterie nicht in einem Messgerät
belassen, das möglicherweise extremen Temperaturen ausgesetzt wird.
9-V-Wechselstromadapter
Für Langzeit-Messwertaufzeichnung kann ein 9-V-
Wechselstromadapter an Stelle der Batterie eingesetzt werden. Ein
solcher Adapter kann unter Berücksichtigung der folgenden
Spezifikationen in einem beliebigen Elektronikgeschäft gekauft werden:
Mini-Netzstecker mit positivem Innenleiter, negativen Außenleiter und
500 mA Nennstromstärke.
Die 9-V-Batterie kann im TMD90 belassen werden, wenn der
Wechselstromadapter verwendet wird.
12
Reparatur
Zu allen Geräten, die zur Reparatur oder Kalibrierung im Rahmen der Garantie
oder außerhalb der Garantie eingesendet werden, muss folgendes beigelegt
werden: Name des Kunden, Firmenname, Adresse, Telefonnummer und
Kaufbeleg. Zusätzlich eine kurze Beschreibung des Problems oder der
gewünschten Wartung sowie die Messleitungen dem Thermometer beilegen.
Die Gebühren für Reparaturen außerhalb der Garantie oder für den Ersatz von
Instrumenten müssen als Scheck, Geldanweisung, Kreditkarte
(Kreditkartennummer mit Ablaufdatum) beglichen werden oder es muss ein
Auftrag an Meterman Test Tools formuliert werden.
Garantiereparaturen oder -austausch – alle Länder
Bitte die nachfolgende Garantieerklärung lesen, und die Batterien und
Sicherungen prüfen, bevor Reparaturen angefordert werden. Während der
Garantieperiode können alle defekten Geräte zum Umtausch gegen dasselbe
oder ein ähnliches Produkt an den Meterman Test Tools-Distributor gesendet
werden. Ein Verzeichnis der zuständigen Distributoren ist im Abschnitt
„Where to Buy“ (Verkaufsstellen) auf der Website
www.metermantesttools.com zu finden. Darüber hinaus können in den USA
und in Kanada Geräte an ein Meterman Test Tools Service-Center (Adresse
siehe weiter unten) zur Reparatur oder zum Umtausch eingesendet werden.
Reparaturen und Austausch außerhalb der Garantie – USA
und Kanada
Für Reparaturen außerhalb der Garantie in den Vereinigten Staaten und in
Kanada werden die Geräte an ein Meterman Test Tools Service-Center
gesendet. Auskunft über die derzeit geltenden Reparatur- und
Austauschgebühren erhalten Sie von Meterman Test Tools oder der
Verkaufsstelle.
In den USA: In Kanada:
Meterman Test Tools Meterman Test Tools
1420 75th Street SW 400 Britannia Rd. E. Unit #1
Everett, WA 98203 Mississauga, ON L4Z 1X9
Tel.: 888-993-5853 Tel.: 905-890-7600
Fax: 425-446-6390 Fax: 905-890-6866
Reparaturen und Austausch außerhalb der Garantie – Europa
Geräte außerhalb der Garantie können durch den zuständigen Meterman Test
Tools-Distributor gegen eine Gebühr ersetzt werden. Ein Verzeichnis der
zuständigen Distributoren ist im Abschnitt „Where to Buy“ (Verkaufsstellen)
auf der Website www.metermantesttools.com zu finden.
Korrespondenzanschrift für Europa*
Meterman Test Tools Europe
P.O. Box 1186
5602 BD Eindhoven
Niederlande
*(Nur Korrespondenz – keine Reparaturen, kein Umtausch unter dieser
Anschrift. Kunden in Europa wenden sich an den zuständigen Distributor.)
13
GARANTIE
Es wird gewährleistet, dass das TMD90 Dual Input Thermometer
innerhalb eines Zeitraums von einem (1) Jahr ab dem Datum des
Kaufes des Thermometers durch den Erstkäufer oder Erstnutzer frei
von Material- oder Fertigungsfehlern ist. Thermometer, die während
der Garantieperiode als defekt angegeben werden, müssen mit dem
Kaufbeleg an ein autorisiertes Meterman Test Tools Service-Center
oder den örtlichen Meterman Test Tools-Händler/-Distributor (bei dem
das Thermometer gekauft wurde) gesendet werden. Nähere
Einzelheiten siehe Abschnitt „Reparatur“. Alle aus dem Kauf eines
Meterman Test Tools-Thermometers abgeleiteten Garantien,
einschließlich, aber nicht beschränkt auf die abgeleiteten Garantien der
Marktfähigkeit oder Eignung für einen bestimmten Zweck, sind auf die
Dauer des oben angegebenen Zeitraums von einem (1) Jahr
beschränkt. Meterman Test Tools haftet nicht für Nutzungsausfall des
Thermometers oder andere Begleit- oder Folgeschäden, Ausgaben oder
wirtschaftliche Verluste oder für jegliche Ansprüche bezüglich solcher
Schäden, Ausgaben oder wirtschaftlicher Verluste. In einigen Ländern
ist die zeitliche Begrenzung der abgeleiteten Gewährleistung sowie der
Ausschluss oder die Begrenzung von Begleit- oder Folgeschäden nicht
zulässig, so dass die oben genannten Einschränkungen und
Ausschlüsse nicht für jeden Käufer gelten. Diese Gewährleistung gibt
dem Eigentümer bestimmte Rechte sowie möglicherweise andere
Rechte, die von Land zu Land verschieden sind.
Thermoelementdefinitionen
Typ Material Material
J
Eisen
Fe
Kupfer-Nickel
(Konstantan)
Cu-Ni
K
Nickel-Chrom
(Chromel)
Ni-Cr
Nickel-Aluminium
Ni-AL
R
Platin-13 % Rhodium
Pt-13% Rh
Platin
Pt
S
Platin-10 % Rhodium
Pt-10% Rh
Platin
Pt
T
Kupfer
Cu
Kupfer-Nickel
(Konstantan)
Cu-Ni
E
Nickel-Chrom
(Chromel)
Ni-Cr
Kupfer-Nickel
(Konstantan)
Cu-Ni
14
Spezifikationen
TE-Typ Bereich Auflösung Genauigkeit
K
-200 bis 650 °C0,1 °C
651 bis 1370 °C1,0 °C
(0,1% Anz. + 0,7 °C)
-328 bis 1000 °F0,1 °F
1001 bis 2498 °F1,0 °F
(0,1% Anz. + 1,4 °F)
J
-200 bis 500 °C0,1 °C
501 bis 760 °C1,0 °C
(0,1% Anz. + 0,7 °C)
-328 bis 940 °F0,1 °C
941 bis 1400 °F1,0 °F
(0,1% Anz. + 1,4 °F
T -200 bis 390 °C0,1 °C (0,1% Anz. + 0,7 °C)
-328 bis 730 °F0,1 °F (0,1% Anz. + 1,4 °F)
R/S 0 bis 1000 °C0,1 °C
1001 bis 1760 °C1,0 °C
(0,3% Anz. + 0,7 °C)
32 bis 1000 °F0,1 °F
1001 bis 3200 °F1,0 °F
(0,1% Anz. + 1,4 °F)
E
-200 bis 380 °C0,1 °C
381 bis 736 °C1,0 °C
(0,3% Anz. + 0,7 °C)
-328 bis 720 °F0,1 °F
721 bis 1832 °F1,0°F
(0,1% Anz. + 1,4 °F)
Weitere Spezifikationen
BETRIEBSBEDINGUNGEN
0 bis 50 °C (32 bis 122 °F) bei 0 bis 85 % relativer Feuchtigkeit (nicht-
kondensierend).
Umgebungskoeffizient
0 bis 18 °C und 28 bis 50 °C (Umgebungstemperatur). Pro °C
Umgebungstemperatur unterhalb von 18 °C bzw. oberhalb von 28 °C
die folgende Toleranz zur Genauigkeitsspezifikation hinzufügen: 0,01 %
Ablesung +0,03 °C (0,01 % Ablesung +0,06 °F)
1
Termometro a due ingressi TMD90
Indice
Informazioni sulla sicurezza..................................................................2
Simboli adoperati nel presente manuale.............................................2
Introduzione........................................................................................3
Display e comandi................................................................................4
Funzioni...............................................................................................5
Spegnimento automatico (modalità di risparmio energetico)..............5
Cambio del tipo di termocoppia.........................................................5
Uso dei pulsanti................................................................................6
Uscita RS-232...................................................................................8
Soluzione dei problemi.........................................................................9
Sostituzione della pila.........................................................................11
Adattatore di c.a. da 9 V.....................................................................11
Riparazioni.........................................................................................12
GARANZIA.........................................................................................13
Descrizione delle termocoppie............................................................13
Dati tecnici......................................................................................14
Dati tecnici aggiuntivi.........................................................................14
Italiano
Italiano
2
Informazioni sulla sicurezza
All’ingresso per termocoppie collegare SOLO termocoppie.
Accertarsi che il termometro sia configurato per l’uso della
termocoppia a disposizione.
Assicurarsi che la termocoppia in uso possa sostenere le
temperature estreme alle quali sarà sottoposta.
Mantenere in modo appropriato il termometro e calibrarlo
regolarmente.
Usare il termometro solo come specificato nel presente manuale,
altrimenti si può compromettere la protezione offerta dallo
strumento.
Sostituire le pile non appena si visualizza il simbolo BAT, per
evitare letture errate che potrebbero causare scosse elettriche e
lesioni.
Avvertenza
La termocoppia fornita non deve entrare in contatto
con liquidi o circuiti elettrici sotto tensione.
Simboli adoperati nel presente manuale
Alta tensione Consultare il manuale
Isolamento doppio Massa di terra
Conforme alle direttive
UE
3
Introduzione
Il modello TMD90 è un termometro a due ingressi che consente il
collegamento di termocoppie di tipo K, J, T, R, S ed E. Dispone di un
display diviso in tre sezioni per la visualizzazione delle misure e
dell’orologio relativo. Il software e il cavo RS-232 opzionali permettono
la trasmissione di dati in tempo reale a un PC per l’ulteriore analisi e
rapporti.
All’accensione, il TMD90 visualizza tutti i segmenti del display a cristalli
liquidi per circa 3 secondi e quindi fornisce informazioni sulle
termocoppie eventualmente collegate. Sono disponibili numerose
configurazioni di visualizzazione. Il display a cristalli liquidi è diviso in
tre sezioni distinte: una più ampia nella parte superiore (visualizzazione
principale) e due più piccole nella parte inferiore (visualizzazione
secondaria e orologio relativo).
Le tre aree di visualizzazione vengono aggiornate continuamente con le
misure della temperatura e le informazioni di tempo relativo. Lo
strumento offre diverse opzioni relative al tipo e al formato dei dati
presenti sul display.
Le letture di temperatura possono essere facilmente convertite da
gradi Fahrenheit a gradi centigradi e viceversa.
La retroilluminazione rende luminoso il display a cristalli liquidi per
una facile lettura in ambienti scarsamente illuminati.
Il pulsante blocca i dati della sezione superiore del display,
mentre la sezione inferiore continua ad aggiornarsi.
L’indicatore di pila scarica è anch’esso aggiornato secondo la
necessità.
Il termometro attiva automaticamente l’ultima modalità selezionata
prima dello spegnimento.
Nota: se non vi sono termocoppie collegate, il visualizzatore della
temperatura presenta quattro trattini (----).
Contenuto della confezione
Termometro TMD90 1
Sonda per termocoppia tipo K 1
Manuale d’uso 1
Custodia di plastica rigida per il trasporto 1
Pila da 9 V (installata) 1
4
Display e comandi
10 (T2)
10 (T1)
5
4
32
6
7
1
8
9
11
6
12
13
REC L’orologio relativo è attivo per la registrazione di
temperature minime e massime o per la marcatura temporale.
MAX/MIN Permette il monitoraggio delle misure di temperatura
massima o minima.
HOLD/REL Blocca il visualizzatore principale o stabilisce lo zero
relativo per i dati in esso presentati.
BAT Indicatore di pila scarica.
°C °F Visualizza la temperatura in gradi Fahrenheit o in gradi
centigradi.
T1-T2 Alterna la visualizzazione come segue: da T1
(principale)/T2 (secondario) a T2 (principale)/T1 (secondario),
quindi a T1-T2 (differenziale) su visualizzatore principale e
alternanza di temperature T1/T2 sul visualizzatore secondario.
Visualizzatore principale. Mostra gli ingressi T1, T2 o T1-T2
(temperatura differenziale - TD) o lo zero relativo di T1, T2 o TD.
KJTRSE Permette di selezionare il riferimento in ingresso corretto
per la termocoppia in uso. T1 e T2 devono essere termocoppie dello
stesso tipo.
Visualizzatore secondario. Indica le temperature T2, T1 o T1 e T2.
Ingressi delle termocoppie. Spinotti polarizzati positivi e negativi
per le sonde delle termocoppie usate (tipo a lama): T1 a sinistra,
T2 a destra.
Visualizzatore dell’orologio relativo. Presenta le ore, i minuti e i
secondi (HHMMSS) quando si preme REC e l’orario relativo in cui
i dati MIN o MAX sono stati registrati.
Uscita RS-232. Software e cavo opzionali (n. parte TM-SW).
Ingresso del convertitore c.a./c.c.
5
Attenzione
Prima di usare il termometro, leggere la sezione
“Informazioni sulla sicurezza”.
Funzioni
Spegnimento automatico (modalità di risparmio
energetico)
Il TMD90 si spegne automaticamente 20 minuti circa dopo
l’accensione. Per registrazioni od operazioni che durano più a lungo, è
possibile disattivare la modalità di risparmio energetico premendo
contemporaneamente i pulsanti e con lo strumento
acceso. Quando al centro dello schermo appare il simbolo “n”, si può
rilasciare il pulsante On. La funzione di spegnimento automatico rimane
disattivata quando si spegne il termometro.
Cambio del tipo di termocoppia
Premere contemporaneamente i pulsanti e per almeno 2
secondi, finché non appare l’indicatore K. Premere il pulsante
per visualizzare uno alla volta i tipi di sonda: K (il tipo predefinito), J, T,
R, S ed E. La modalità corrente appare sul lato sinistro del display a
cristalli liquidi. Premere : al centro dello schermo appare una
S”.
6
Uso dei pulsanti
Il display visualizza automaticamente l’ultima modalità selezionata.
Inoltre, per comodità, il termometro richiama le impostazioni usate
durante l’ultima operazione.
˚C ˚F
HOLD
REL
REC
T
1 T2
7
Usare i pulsanti per comandare le varie funzioni del termometro
TMD90. Nella tabella si presume quanto segue: il TMD90 è stato
acceso, sono state collegate due termocoppie, il visualizzatore
principale presenta la termocoppia T1 (predefinita), la termocoppia è di
tipo K e la registrazione è disattivata.
Pulsante
Descrizione
Accende e spegne il termometro.
Premendo brevemente il pulsante REL/HOLD, il
visualizzatore principale (T1, T2 o T1-T2) si blocca e
sopra di esso appare l’indicatore HOLD; premendolo per
2 secondi o più, nella parte superiore del display a
cristalli liquidi appare l'indicatore REL e il visualizzatore
principale (REL) presenta lo zero relativo. Lo zero
relativo porta a "000.0" il valore nel visualizzatore
principale; dopodiché, il display indica solo le variazioni
di temperatura. In questa modalità è possibile registrare
le temperature relative. Premere di nuovo il pulsante
REL/HOLD per riportare lo strumento all'impostazione
predefinita.
Premendo brevemente il pulsante, il visualizzatore
principale passa all’ingresso T2 (lo schermo secondario
visualizza T1). Premendolo di nuovo brevemente, sul
visualizzatore principale appare T1-T2 (quello
secondario alterna tra T1 e T2). Premendolo una terza
volta, lo strumento ritorna all’impostazione predefinita
.
Premendo brevemente il pulsante, la retroilluminazione
si accende per circa 30 secondi e poi si spegne.
Premendo brevemente il pulsante, il display da gradi
Fahrenheit (unità predefinita) a gradi centigradi e
viceversa. L’unità corrente è indicata sul lato destro del
display.
Premendo brevemente il pulsante, nell’angolo inferiore
destro del display si avvia l’orologio relativo, mentre
nell’angolo superiore sinistro si accende l’indicatore. Tutte
le altre funzioni sono bloccate, ad eccezione dei pulsanti
di alimentazione e di retroilluminazione. Il visualizzatore
principale indica T1, T2 o T1-T2, mentre quello
secondario continua ad aggiornarsi. Premendolo di nuovo
brevemente, lo strumento passa dalla visualizzazione della
temperatura massima registrata (MAX) a quella minima
(MIN) e poi di nuovo a quella corrente. Il display
visualizza la modalità di registrazione. Per disattivarla,
tenere premuto il pulsante per tre
secondi.
8
Uscita RS-232
Con il software e il cavo RS-232 opzionali (n. parte T90/10-SW), il
TMD90 è in grado di inviare i risultati delle misure a un computer con
sistema operativo Windows 95/98/NT/2000/XP/ME. Tra le
caratteristiche del software, vi sono:
registrazione di un massimo di 16.000 letture campione con
memoria dell’orologio di tempo reale;
intervallo di campionamento programmabile da 1 secondo a 3.600
secondi;
limiti di allarme minimo e massimo definiti dall’utente;
possibilità di analisi ulteriori mediante lo scaricamento di dati in un
programma del computer;
memorizzazione di tutti i dati nel formato di testo (.txt) per
agevolarne l’importazione in un altro programma software.
9
Soluzione dei problemi
Problema Soluzione
Il termometro non si accende.
Controllare la tensione della
pila e sostituirla se è
scarica.
Verificare che i terminali
della pila facciano un buon
contatto con i morsetti.
I visualizzatori dei dati T1 e T2
presentano dei trattini.
Collegare la termocoppia
mancante.
I visualizzatori dei dati T1 e/o
T2 presentano dei trattini
benché le termocoppie siano
collegate.
Misurare la resistenza delle
termocoppie per accertarsi
che non vi siano dei guasti
interni.
Pulire le termocoppie e
riprovare.
10
Problema Soluzione
Si verifica una deriva della
temperatura da un valore noto
in un ambiente controllato.
Verificare che il tipo di
termocoppia corrisponda
all’icona visualizzata.
Pulire e asciugare le lame
delle termocoppie e lasciare
che si asciughino all’aria.
Verificare la temperatura
con una termocoppia
funzionante. Sostituire la
termocoppia eventualmente
difettosa.
L’orologio relativo non parte
quando si preme il
pulsante .
Verificare che la
termocoppia si stata
inserita bene.
Appaiono dei trattini durante la
revisione del valore massimo
registrato.
La termocoppia non è stata
riconosciuta. Verificare che
non vi sia un contatto
intermittente o che la
termocoppia non sia stata
rimossa temporaneamente.
I dati continuano ad
aggiornarsi anche dopo aver
premuto il pulsante .
Verificare la presenza
dell'icona HOLD sul
display. Premere il pulsante
con decisione.
Lo strumento si spegne
durante la registrazione.
Premere
contemporaneamente i
pulsanti e
prima di accendere lo
strumento per disattivare la
funzione di spegnimento
automatico.
11
Sostituzione della pila
Sostituire la pila da 9 V quando:
l’icona BAT appare nell’angolo destro del display;
il termometro non si accende;
l’icona BAT appare quando si accende la retroilluminazione.
Se il termometro non risponde ai comandi, controllare la tensione della
pila, anche se è stata sostituita di recente.
Sostituzione della pila
1. Scollegare tutte le termocoppie dalla parte superiore del
termometro.
2. Appoggiare lo strumento su una superficie pulita e in piano, con la
parte anteriore rivolta verso il basso.
3. Rimuovere lo sportello dello scomparto della pila, dopo averne
allentato la vite.
4. Sostituire la batteria.
5. Rimontare lo sportello e serrare la vite.
Se si prevede di non usare il termometro per un mese o più, rimuovere
la pila. Rimuoverla anche nel caso in cui lo strumento venga esposto a
temperature estreme.
Adattatore di c.a. da 9 V
Per la registrazione a lungo termine delle misure, anziché alimentare lo
strumento a batteria si può usare un adattatore di c.a. a 9 V. Questo
accessorio può essere acquistato in un qualsiasi negozio di componenti
elettronici, ma deve soddisfare i seguenti requisiti: spinotto di
alimentazione mini-plug con conduttore interno positivo e conduttore
esterno negativo, 500 mA di corrente nominale. La pila da 9 V può
essere lasciata nel TMD90 durante l’uso dell’adattatore.
12
Riparazioni
A tutti gli strumenti di misura restituiti per interventi in garanzia o non
coperti dalla garanzia, oppure per la taratura, devono essere allegate le
seguenti informazioni: il proprio nome e quello dell’azienda, indirizzo,
numero telefonico e scontrino. Allegare anche una breve descrizione
del problema o dell’intervento richiesto e i cavetti. Gli importi dovuti per
sostituzioni o riparazioni non coperte dalla garanzia vanno versati
tramite assegno, vaglia bancario, carta di credito con data di scadenza
od ordine di acquisto all’ordine di Meterman Test Tools.
Sostituzioni e riparazioni in garanzia – Tutti i Paesi
Si prega di leggere la seguente garanzia e di controllare la pila e i
fusibili prima di richiedere una riparazione. Durante il periodo di
garanzia, si può restituire uno strumento difettoso al rivenditore
Meterman Test Tools per ricevere un prodotto identico o analogo. Nella
sezione “Where to Buy” del sito c’è un elenco dei distributori più vicini.
Negli Stati Uniti e nel Canada gli strumenti da sostituire o riparare in
garanzia possono essere inviati anche a un centro di assistenza
Meterman Test Tools (l’indirizzo è più avanti).
Sostituzioni e riparazioni non coperte dalla garanzia – Usa e
Canada
Per riparazioni non coperte dalla garanzia, negli Stati Uniti e nel Canada
lo strumento deve essere inviato a un centro di assistenza Meterman
Test Tools. Rivolgersi alla Meterman Test Tools o al rivenditore per
informazioni sui costi delle riparazioni e sostituzioni.
USA Canada
Meterman Test Tools Meterman Test Tools
1420 75th Street SW 400 Britannia Rd. E. Unit #1
Everett, WA 98203 Mississauga, ON L4Z 1X9
Tel: 888 993 5853 Tel: 905 890 7600
Fax: 425 446 6390 Fax: 905 890 6866
Sostituzioni e riparazioni non coperte dalla garanzia – Europa
Gli strumenti acquistati in Europa e non coperti dalla garanzia possono
essere sostituiti dal rivenditore Meterman Test Tools per un importo
nominale. Nella sezione “Where to Buy” del sito c’è un elenco dei
distributori più vicini.
Recapito postale europeo*
Meterman Test Tools Europe
P.O. Box 1186
5602 BD Eindhoven
Paesi Bassi
*(Solo per corrispondenza – non rivolgersi a questo indirizzo per
riparazioni o sostituzioni. Si pregano i clienti europei di rivolgersi al
rivenditore.)
13
GARANZIA
Si garantisce che il termometro a due ingressi TMD90 sarà esente da
difetti di materiale e di fabbricazione per un anno a decorrere dalla data
di acquisto da parte dell’acquirente o dell’utente originale. Se durante il
periodo di garanzia si ritiene che il termometro sia difettoso, restituirlo,
allegando lo scontrino, a un centro di assistenza Meterman Test Tools
oppure al rivenditore o distributore locale Meterman Test Tools presso
cui è stato acquistato. Per ulteriori informazioni vedere la sezione
Riparazioni. La durata di qualsiasi garanzia implicita attivata in base alla
vendita di un termometro Meterman Test Tools, incluse ma non a titolo
esclusivo le garanzie implicite di commerciabilità e idoneità per uno
scopo specifico, è limitata al periodo sopra specificato di un anno. La
Meterman Test Tools non sarà responsabile del mancato utilizzo del
termometro o di altri danni incidentali o indiretti, di spese o perdite
economiche, o delle relative richieste di risarcimento. Alcuni Paesi non
consentono limitazioni sulla durata delle garanzie implicite e/o
l’esclusione o la limitazione di danni incidentali o indiretti; di
conseguenza le precedenti limitazioni o esclusioni non saranno
applicate. Questa garanzia offre specifici diritti legali ed è possibile che,
secondo il Paese in cui si vive, si abbiano altri diritti.
Descrizione delle termocoppie
Tipo Lega Lega
J
Ferro
Fe
Rame-nichel
(Constantan)
Cu-Ni
K
Nichel-cromo
(Chromel)
Ni-Cr
Nichel-alluminio
Ni-AL
R
Platino-rodio al 13%
Pt-13% Rh
Platino
Pt
S
Platino-rodio al 10%
Pt-10% Rh
Platino
Pt
T
Rame
Cu
Rame-nichel
(Constantan)
Cu-Ni
E
Nichel-cromo
(Chromel)
Ni-Cr
Rame-nichel
(Constantan)
Cu-Ni
14
Dati tecnici
Tipo T/C Portata Risoluzione Precisione
K
da -200 a 650 °C0,1 °C
da 651 a 1370 °C1,0 °C
(0,1% lettura + 0,7 °C)
da -328 a 1000 °F0,1 °F
da 1001 a 2498
°F
1,0 °F
(0,1% lettura +1,4 °F)
J
da -200 a 500 °C0,1 °C
da 501 a 760 °C1,0 °C
(0,1% lettura + 0,7 °C)
da -328 a 940 °F0,1 °C
da 941 a 1400 °F1,0 °F
(0,1% lettura + 1,4 °F)
T da -200 a 390 °C0,1 °C
(0,1% lettura + 0,7 °C)
da -328 a 730 °F0,1 °F
(0,1% lettura +1,4 °F)
R/S da 0 a 1000 °C0,1 °C
da 1001 a 1760
°C
1,0 °C
(0,3% lettura + 0,7 °C)
da 32 a 1000 °F0,1 °F
da 1001 a 3200
°F
1,0 °F
(0,1% lettura +1,4 °F)
E
da -200 a 380 °C0,1 °C
da 381 a 736 °C1,0 °C
(0,3% lettura + 0,7 °C)
da -328 a 720 °F0,1 °F
da 721 a 1832 °F1,0°F
(0,1% lettura +1,4 °F)
Dati tecnici aggiuntivi
CONDIZIONI DI FUNZIONAMENTO
Da 0 a 50 °C (32-122 °F) a umidità relative fra 0 e 85% (senza
condensa)
Coefficiente ambientale
Da 0 a 18 °C e da 28 a 50 °C (temperature ambiente). Per ciascun
grado centigrado di temperatura ambiente al di sotto di 18 °C o sopra
28 °C, aggiungere nelle specifiche di precisione la seguente tolleranza:
0,01% della lettura + 0,03 °C (0,01% della lettura + 0,06 °F).
1
Termómetro TMD90 de doble entrada
Índice
Información sobre seguridad................................................................2
Símbolos utilizados en este manual...................................................2
Introducción........................................................................................3
Pantalla y controles..............................................................................4
Funciones............................................................................................5
Apagado automático (modo de reposo).............................................5
Cambio del tipo de termopar.............................................................5
Uso de los botones pulsadores..........................................................6
Salida RS-232...................................................................................8
Localización y resolución de fallos........................................................9
Reemplazo de la batería .....................................................................11
Adaptador de 9 V CA..........................................................................11
Reparación........................................................................................12
GARANTÍA.........................................................................................13
Definiciones de los termopares...........................................................13
Especificaciones..............................................................................14
Especificaciones adicionales...............................................................14
Español
Español
2
Información sobre seguridad
Conecte ÚNICAMENTE termopares en la entrada de termopares.
Asegúrese de que su medidor esté configurado para el tipo de
termopar que se utilizará.
Cerciórese de que el termopar que utilice pueda soportar la
temperatura extrema a la cual estará expuesto.
Mantenga adecuadamente el medidor y calíbrelo con regularidad
Utilice el termómetro solamente de la manera que se especifica en
este manual; de lo contrario, la protección provista por el
termómetro podría verse afectada.
Reemplace la batería tan pronto como aparezca BAT para evitar
lecturas falsas que podrían traer como consecuencia choques
eléctricos y lesiones.
Advertencia
El termopar suministrado no está diseñado para tener
contacto con líquidos ni circuitos eléctricos con
presencia de tensión.
Símbolos utilizados en este manual
Tensión peligrosa Consulte el manual
Aislamiento doble Conexión a tierra
Cumple con las
directivas de la Unión
Europea.
3
Introducción
El modelo TMD90 es un termómetro de doble entrada que acepta
termopares tipo K, J, T, R, S y E. Cuenta con una pantalla de
visualización triple para ver los resultados de la medición, además de
un reloj de tiempo relativo. El software y cable RS-232 opcionales
permiten la recolección de datos en tiempo real en un PC, para su
posterior análisis o redacción de informes.
El modelo TMD90 muestra todos los segmentos LCD (pantalla de
cristal líquido) durante aproximadamente 3 segundos cuando se lo
enciende inicialmente, y luego proporciona información sobre cualquier
termopar que se hubiera conectado. Se dispone de numerosas
combinaciones de visualización. La pantalla LCD está dividida en tres
secciones distintas: una pantalla superior grande (principal), y dos
pantallas inferiores más pequeñas (secundaria y reloj relativo).
Las tres zonas de visualización se actualizan continuamente con las
mediciones de temperatura y la información sobre tiempo relativo.
Existen diversas opciones referentes a cómo y cuál información se
presenta en la pantalla LCD.
Las lecturas de temperatura se alternan fácilmente entre grados
Fahrenheit y centígrados.
Una luz de fondo ilumina la pantalla LCD para permitir la
visualización en áreas de luz tenue.
El botón congelará los datos de la pantalla superior mientras
permite al mismo tiempo que las pantallas inferiores pueden
continuar actualizando la información.
También se exhibe un indicador de batería con poca carga cuando
corresponda.
El medidor exhibe de manera predeterminada el último modo
seleccionado cuando se lo había apagado.
Nota: Si no hay termopares conectados, aparecerán cuatro guiones (---
-) en la pantalla de temperatura.
Equipo incluido
Termómetro TMD90 1
Sonda de termopar tipo K 1
Manual del usuario 1
Estuche de transporte de plástico duro 1
Batería de 9 V (instalada) 1
4
Pantalla y controles
10 (T2)
10 (T1)
5
4
32
6
7
1
8
9
11
6
12
13
REC El reloj relativo está activo para registrar las ocurrencias de
temperatura mínima y máxima o para mostrar fecha y hora.
MAX/MIN Se están supervisando las mediciones de temperatura
mínima o máxima.
HOLD/REL Congela la pantalla primaria o establece un cero
relativo para la información de la pantalla principal.
BAT Indicador de carga escasa de la batería.
°C °F Muestra la temperatura en grados centígrados o
Fahrenheit.
T1-T2 Alterna la información de las pantallas desde T1
(principal)/ T 2 (secundaria) a T2 (principal) / T1 (secundaria),
luego de T 1-T 2 (diferencial) en la pantalla principal y alternando
las temperaturas T1/T2 en la pantalla secundaria.
Pantalla principal de datos. Muestra T1, T2 o T1-T2 (diferencial
de temperatura -TD ) o un cero relativo de T1, T2 o TD.
KJTRSE Selecciona la referencia de entrada correcta para el
termopar en uso. T1 y T2 deben ser del mismo tipo de termopar.
Pantalla secundaria de datos. Muestra T2, T1 o las temperaturas T1 y T2.
Entrada del termopar. Enchufes polarizados positivo y negativo
para las sondas del termopar utilizadas (tipo de patilla); T1 a la
izquierda, T2 a la derecha.
Pantalla de reloj relativo. Exhibe el tiempo en horas, minutos y
segundos (HHMMSS) al pulsarse REC, y el tiempo relativo en
que fueron registrados los datos MIN o MAX.
Puerto de salida RS-232. Software y cable opcionales – PN TM-
SW.
Entrada del convertidor de CA a CC.
5
Precaución
Lea toda la información de seguridad antes de utilizar
este termómetro.
Funciones
Apagado automático (modo de reposo)
El medidor TMD90 se apaga automáticamente aproximadamente
20 minutos después de ser encendido. Para registrar u operar a lo
largo de períodos de tiempo más prolongados, puede desactivar el
modo de reposo pulsando y simultáneamente, cuando se
enciende el instrumento. Cuando aparece "n" en el centro de la pantalla,
puede soltar el botón On. Se desactiva la función de apagado
automático al apagarse el medidor.
Cambio del tipo de termopar
Pulse y simultáneamente durante 2 o más segundos
hasta que aparezca K. Pulse el botón . El tipo de sonda pasará
cíclicamente a través de los tipos K (el tipo predeterminado), J, T, R, S
y E. El modo vigente se muestra del lado izquierdo de la pantalla LCD.
Pulse ; aparecerá una "S" en el centro de la pantalla.
6
Uso de los botones pulsadores
La pantalla arranca de manera predeterminada en el último modo
utilizado. Para su conveniencia, el medidor utiliza de manera
predeterminada los ajustes empleados durante la última operación.
˚C ˚F
HOLD
REL
REC
T
1 T2
7
Para controlar la operación del TMD 90 utilice los pulsadores. La tabla supone
que se ha encendido el TMD90 con dos termopares instalados y que se lo ha
configurado de modo de mostrar T1 (opción predeterminada) en la pantalla
principal, un termopar de tipo K y la función de registro apagada.
Botón Descripción
Enciende y apaga el medidor.
Pulse el botón HOLD/REL momentáneamente y la
pantalla principal (T1, T2 o T1-T2) se congela, con
la palabra HOLD exhibida en la parte superior. Pulse
este botón durante dos o más segundos y aparecerá
REL en la parte superior de la pantalla LCD; la
pantalla principal REL indica el cero relativo. El cero
relativo hace que el valor de la pantalla principal se
muestre como "000.0" – sólo la medida del cambio
de temperatura será indicada. Se pueden registrar
temperaturas relativas. Pulse HOLD/REL una vez
más para que la unidad regresa a su valor
predeterminado.
Pulse momentáneamente y la pantalla principal cambiará
a T2 (la pantalla secundaria muestra T1); pulse este botón
mo-mentáneamente una vez más y mostrará T1-T2; la
pantalla secundaria alterna entre T1 y T2; pulse
momentáneamente una vez más y el instrumento retornará a
su valor predeterminado.
Pulse este botón momentáneamente y la luz de
fondo se iluminará durante aproximadamente 30
segundos, para luego apagarse.
Pulse este botón momentáneamente y la unidad alternará
entre las temperaturas en Fahrenheit (la opción
predeterminada) y en grados centígrados. El modo
vigente está indicado del lado derecho de la pantalla LCD.
Pulse este botón momentáneamente y el reloj relativo
comenzará a funcionar en la pantalla inferior derecha;
aparecerá del lado izquierdo superior de la pantalla LCD –
Todas las demás funciones del botón quedarán
bloqueadas, salvo Power y Backlight. T1, T2 o T1-T2 se
muestran en la pantalla primariaprincipal; la pantalla
secundaria continúa actualizándose. Pulse
momentáneamente una vez más y la unidad pasará
cíclicamente a través de las temperaturas MAX y MIN
(temperaturas máxima y mínima registardas) y regresará
a la temperatura actual. El modo de registro se muestra en
la pantalla LCD. Pulse y mantenga pulsado este botón
durante tres segundos para desactivar la función de
registro.
8
Salida RS-232
Con el software y cable opcionales, P/N TM-SW, el TMD90 puede
enviar los resultados de las mediciones a una computadora que cuente
con alguno de los siguientes sistemas operativos: Windows
95/98/NT/2000/XP/ME. Características del software:
Registra hasta 16,000 lecturas de muestra con memoria del reloj en
tiempo real.
Intervalo de muestras programable desde 1 seg hasta 3600 seg.
Configuraciones de alarma máxima y mínima definidos por el
usuario.
Permite un posterior análisis mediante la descarga de los datos a
un programa informático.
Todos los datos se guardan en formato .txt para su fácil
transformación en otro programa de software.
9
Localización y resolución de fallos
Problema Solución
El termómetro no se enciende.
Verifique el voltaje de la
batería y reemplace la
misma si está descargada.
Verifique que los clips de la
batería sujeten de manera
firme los bornes de la
misma.
Aparecen guiones en las
pantallas de datos T1 y T2.
Inserte el termopar faltante.
Aparecen guiones en las
pantallas de datos T1 y/o T2,
con los termopares
insertados.
Mida la resistencia de los
termopares para
asegurarse de que no estén
partidos internamente.
Limpie el termopar y
reinicie el proceso.
10
Problema Solución
Tienen lugar desplazamientos
de temperatura desde un valor
conocido en un entorno
controlado
Verifique que el tipo de
termopar coincida con el
icono mostrado.
Limpie y seque las patillas
del termopar y permita que
se sequen al aire.
Confirme la temperatura
con un termopar que se
sepa que está en buenas
condiciones. En caso
necesario reemplace el
termopar defectuoso.
El reloj relativo no se inicia
cuando se pulsa el botón
.
Verifique que el termopar
se haya insertado
correctamente.
Aparecen guiones durante una
revisión del valor máximo
registrado.
El termopar no se
reconoce. Compruebe que
no haya una desconexión
intermitente o momentánea
del termopar.
Los datos continúan
actualizándose después de
que usted pulsa el botón
.
Verifique que aparezca el
icono HOLD en la pantalla.
Pulse el botón con
firmeza.
El instrumento se apaga
mientras se encuentra
registrando.
Pulse y
simultáneamente antes del
encendido para desactivar
la característica de apagado
automático.
11
Reemplazo de la batería
Reemplace la batería de 9 V cuando:
Aparece BAT del lado derecho de la pantalla.
El medidor no se enciende.
Aparece BAT cuando usted enciende la luz de fondo.
Incluso si se reemplazó la batería recientemente, verifique el nivel de
voltaje en caso de no haber respuesta del medidor.
Para reemplazar la batería
1. Retire todos los termopares de la parte superior del medidor.
2. Coloque el instrumento boca abajo sobre una superficie plana y
limpia.
3. Afloje el tornillo y retire la puerta del compartimiento de la batería.
4. Reemplace la batería.
5. Instale nuevamente la puerta del compartimiento de la batería y
apriete el tornillo.
Retire la batería si no tiene intención de utilizar el medidor durante un
mes o más. No deje la batería en un medidor que pueda ser expuesto a
extremos de temperatura.
Adaptador de 9 V CA
Para el registro de mediciones a largo plazo, se puede usar un
adaptador de 9 V CA en lugar de la batería. Dicho adaptador se puede
adquirir en cualquier tienda de artículos electrónicos, ajustándose a las
siguientes especificaciones: Enchufe eléctrico en miniatura con positivo
y conductor externo negativo, con un valor nominal de corriente de 500
mA. Cuando se utiliza el adaptador de CA, la batería de 9V puede
permanecer en el TMD90.
12
Reparación
Todas las herramientas de prueba devueltas para su calibración o reparación, cubiertas
o no por la garantía, deberán estar acompañadas por lo siguiente: su nombre, el
nombre de la empresa, la dirección, el número de teléfono y una prueba de compra.
Además, incluya una breve descripción del problema o del servicio solicitado y las
puntas de prueba del termómetro. Los pagos correspondientes a reparaciones o
reemplazos no cubiertos por la garantía se deben remitir a la orden de Meterman Test
Tools en forma de cheque, giro postal, pago mediante tarjeta de crédito (incluir el
número y la fecha de vencimiento) u orden de compra.
Reparaciones y reemplazos cubiertos por la garantía (todos
los países)
Antes de solicitar una reparación sírvase leer la siguiente declaración de garantía y
compruebe el estado de las baterías y los fusibles de su medidor. Durante el periodo de
garantía, toda herramienta de comprobación en mal estado de funcionamiento puede
ser devuelta al distribuidor de Meterman Test Tools para cambiarla por otra igual o un
producto semejante. Consulte la sección “Dónde comprar” del sitio
www.metermantesttools.com en Internet para obtener una lista de los distribuidores de
su zona. Además, en los Estados Unidos y Canadá, las unidades para reparación y
reemplazo cubiertas por la garantía también se pueden enviar a un Centro de Servicio
de Meterman Test Tools (las direcciones se incluyen más adelante).
Reparaciones y reemplazos no cubiertos por la garantía
(Estados Unidos y Canadá)
Las unidades para reparaciones no cubiertas por la garantía en Estados Unidos y
Canadá se deben enviar a un Centro de Servicio de Meterman Test Tools. Póngase en
contacto con Meterman Test Tools o con el vendedor de su producto para solicitar
información acerca de los precios vigentes para reparación y reemplazo.
En Estados Unidos En Canadá
Meterman Test Tools Meterman Test Tools
1420 75th Street SW 400 Britannia Rd. E. Unit #1
Everett, WA 98203 Mississauga, Ontario L4Z 1X9
Tel.: 888-993-5853 Tel.: 905-890-7600
Fax: 425-446-6390 Fax: 905-890-6866
Reparaciones y reemplazos no cubiertos por la garantía –
Europa
El distribuidor de Meterman Test Tools puede reemplazar aplicando un cargo nominal
las unidades vendidas en Europa no cubiertas por la garantía. Consulte la sección
“Dónde comprar” del sitio www.metermantesttools.com en Internet para obtener una
lista de los distribuidores de su zona.
Dirección para envío de correspondencia en Europa*
Meterman Test Tools Europe
P.O. Box 1186
5602 B.D. Eindhoven
Holanda
*(Correspondencia solamente. En esta dirección no se suministran reparaciones ni
reemplazos. Los clientes europeos deben ponerse en contacto con el distribuidor).
13
GARANTÍA
El termómetro TMD90 Dual Input está garantizado contra defectos en
los materiales o la mano de obra por un periodo de un (1) año a partir
de la fecha de compra del mismo por el comprador o usuario original.
Todo termómetro considerado defectuoso durante el periodo de
garantía debe ser devuelto con la correspondiente prueba de compra a
un Centro de Servicio autorizado de Meterman Test Tools o al
representante o distribuidor local en donde fue adquirido el
instrumento. Consulte la sección Reparación para obtener información
más detallada. La duración de todas las garantías implícitas resultantes
de la venta de un termómetro Meterman Test Tools, incluidas pero no
limitadas a las garantías de comerciabilidad e idoneidad para un
propósito determinado, se limitan al periodo de garantía de un (1) año
establecido anteriormente. Meterman Test Tools no se responsabiliza
por la imposibilidad de utilizar el termómetro u otros perjuicios, gastos
o pérdidas económicas consecuentes o incidentales, ni por ningún tipo
de reclamo o reclamos por tales perjuicios, gastos o pérdidas
económicas. Ya que algunos estados no permiten la limitación de la
duración de una garantía implícita, ni la exclusión ni limitación de
daños incidentales o consecuentes, las limitaciones y exclusiones
anteriores pueden no regir para usted. Esta garantía le otorga derechos
legales específicos y es posible que también tenga otros derechos que
varían de un estado a otro.
Definiciones de los termopares
Tipo Aleación Aleación
J
Hierro
Fe
Cobre-Níquel
(Constantan)
Cu-Ni
K
Níquel-Cromo (Chromel)
Ni-Cr
Níquel-Aluminio
Ni-AL
R
Platino-13% Rodio
Pt-13% Rh
Platino
Pt
S
Platino-10% Rodio
Pt-10% Rh
Platino
Pt
T
Cobre
Cu
Cobre-Níquel
(Constantan)
Cu-Ni
E
Níquel-Cromo (Chromel)
Ni-Cr
Cobre-Níquel
(Constantan)
Cu-Ni
14
Especificaciones
Tipo de
termopar
Rango Resolución Precisión
K
-200 a 650 °C0,1 °C
651 a 1370 °C1,0 °C
(0,1% lectura + 0,7 °C)
-328 a 1000 °F0,1 °F
1001 a 2498 °F1,0 °F
(0,1% lectura + 1,4 °F)
J
-200 a 500 °C0,1 °C
501 a 760 °C1,0 °C
(0,1% lectura + 0,7 °C)
-328 a 940 °F0,1 °C
941 a 1400 °F1,0 °F
(0,1% lectura + 1,4 °F
T
-200 a 390 °C0,1 °C (0,1% lectura + 0,7 °C)
-328 a 730 °F0,1 °F (0,1% lectura + 1,4 °F)
R/S 0 a 1000 °C0,1 °C
1001 a 1760 °C1,0 °C
(0,3% lectura + 0,7 °C)
32 a 1000 °F0,1 °F
1001 a 3200 °F1,0 °F
(0,1% lectura + 1,4 °F)
E
-200 a 380 °C0,1 °C
381 a 736 °C1,0 °C
(0,3% lectura + 0,7 °C)
-328 a 720 °F0,1 °F
721 a 1832 °F1,0°F
(0,1% lectura + 1,4 °F)
Especificaciones adicionales
CONDICIONES DE OPERACIÓN
0 a 50 °C (32 a 122 °F) a valores de humedad relativa del 0 al 85 %
(sin condensación)
Coeficiente ambiente
0 a 18 °C y 28 a 50 °C (temperaturas ambiente)
Para cada °C de temperatura ambiente menor que 18 °C o mayor que
28 °C, agregue la siguiente tolerancia a la especificación de exactitud:
0,01% de la lectura + 0,03 °C (0,01% de la lectura + 0,06 °F)
TMD90TMD90
Dual Input
Digital
Thermometer
User Manual
Mode d'emploi
Bedienungshandbuch
Manuale d'Uso
Manual de uso
U.S. Service Center
Meterman Test Tools
1420 75th Street SW
Everett, WA 98203
Tel: 888-993-5853
Fax: 425-446-6390
Canadian Service Center
Meterman Test Tools
400 Britannia Rd. E. Unit #1
Mississauga, ON L4Z 1X9
Tel: 905-890-7600
Fax: 905-89-6866
European Correspondence Address*
Meterman Test Tools Europe
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
*Correspondence only - no repair or replacement
available from this address. European customers
please contact your distributor.
Visit www.metermantesttools.com for
• Catalog
• Application notes
• Product specifications
• Product manuals
PN 2099366
September 2003
© Wavetek Meterman Test Tools.
All rights reserved. Printed in China.
Please Recycle
®
®
®
®

Transcripción de documentos

TMD90 TMD90 ® ® ® ® U.S. Service Center Canadian Service Center Dual Input Digital Thermometer European Correspondence Address* User Manual Meterman Test Tools 1420 75th Street SW Everett, WA 98203 Tel: 888-993-5853 Fax: 425-446-6390 Meterman Test Tools 400 Britannia Rd. E. Unit #1 Mississauga, ON L4Z 1X9 Tel: 905-890-7600 Fax: 905-89-6866 Meterman Test Tools Europe P.O. Box 1186 5602 BD Eindhoven The Netherlands *Correspondence only - no repair or replacement available from this address. European customers please contact your distributor. Visit www.metermantesttools.com for • Catalog • Application notes • Product specifications • Product manuals PN 2099366 September 2003 © Wavetek Meterman Test Tools. All rights reserved. Printed in China. Please Recycle • Mode d'emploi • Bedienungshandbuch • Manuale d'Uso • Manual de uso Document Dimensions: 3 x 5 inches TMD90 Dual Input Thermometer Safety Information .............................................................................. 2 Symbols Used in this Manual ........................................................... 2 Introduction........................................................................................ 3 Display and Controls ........................................................................... 4 Functions............................................................................................ 5 Auto Power Off ( Sleep Mode ) ......................................................... 5 Change Thermocouple Type ............................................................. 5 Using the Pushbuttons..................................................................... 6 RS-232 Output................................................................................. 7 Troubleshooting.................................................................................. 8 Replacing the Battery .......................................................................... 9 9 V ac Adapter .................................................................................... 9 Repair............................................................................................... 10 WARRANTY...................................................................................... 11 Thermocouple Definitions ................................................................. 11 Specifications ................................................................................... 12 Additional Specifications ................................................................... 12 1 English Contents Safety Information • • • • • • Place ONLY thermocouples in the thermocouple input. Make sure your meter is configured for the thermocouple type to be used. Be sure the thermocouple you use can withstand the temperature extreme it will be exposed to. Properly maintain the meter and calibrate it regularly Use the thermometer only as specified in this manual, or the protection provided by the thermometer might be impaired. Replace the battery as soon as BAT appears to avoid false readings that can lead to electric shock and injury. Warning Supplied thermocouple is not intended for contact with liquids or live electrical circuits. XW Symbols Used in this Manual X T P W J Dangerous Voltage Double insulated Complies with EU directives 2 Refer to the manual Earth Ground Introduction The Model TMD90 dual input thermometer accepts Type K, J, T, R, S, and E thermocouples. With a triple display screen to view measurement results and relative time clock. Optional RS-232 software and cable allows collection of real time data to a PC for further analysis or reports. TMD90 displays all LCD (liquid crystal display) segments for approximately 3 seconds when it is first turned on and then provides information on any connected thermocouples. Numerous viewing combinations are available. The LCD is divided into three distinct sections; one large (Primary) top screen and two smaller bottom screens (Secondary and Relative Clock). The three display areas are continually updated with the temperature measurements and relative time information. There are several options regarding how and what information is presented on the LCD. • Temperature readings are easily toggled between Fahrenheit and Celsius. • A backlight illuminates the LCD for viewing in low light areas. button will freeze the upper display data while allowing • The the lower displays to continue updating Information. • A low battery indicator is also displayed as appropriate. • The meter defaults to the last mode selected when turned off. Note: If no thermocouples are connected, four dashes(----) appear in the temperature display. Box Contents TMD90 Thermometer 1 Type K thermocouple probe 1 Users Manual 1 Hard plastic carry case 1 9 V battery (installed) 1 A 3 Display and Controls 10 (T2) 10 (T1) 2 3 4 1 5 7 6 8 11 9 6 12 13 A B C D E F G H I J K L M REC Relative clock is active for recording min and max temperature occurences or for time stamp purposes. MAX/MIN Maximum or minimum temperature measurements are being monitored. HOLD/REL Freezes primary display or establishes a relative zero for the primary display information. BAT Low battery indicator. °C °F Displays temperature in either degrees Celsius or degrees Fahrenheit. T1-T2 Toggles screen niformation from T1 (Primary)/ T 2 (Secondary) to T2 (Primary) / T1 (Secondary) ,then to T 1-T 2 (Differential) on Primary and alternating T1/T2 temperatures on Secondary display. Primary data display. Displays T1, T2, or T1-T2( temperature differential -TD ) or a relative zero of T1, T2, or TD. KJTRSE Selects proper input reference for thermocouple in use. T1 and T2 must be the same thermocouple type. Secondary data display. Displays T2,T1 or T1 and T2 temperatures. Thermocouple input. Positive and negative polarized plugs for the thermocouple probes used ( Blade type ); T1 on the left ,T2 on the right. Relative clock display. Displays time in hours, minutes, and seconds (HHMMSS) when REC is pressed and the relative time that MIN or MAX data was recorded. RS-232 output port. Optional software and cable – PN TM-SW. AC to DC converter input. 4 W Caution Read all Safety Information before using this thermometer. Functions Auto Power Off (Sleep Mode) DA The TMD90 shuts off automatically approximately 20 minutes after being turned on. For recording or operating over longer periods of time you can disable the sleep mode by pressing and simultaneously while power on. When "n" then appear in the center of the screen you can release the On. Auto power off is disabled when you turn off the meter. E DC E Change Thermocouple Type Press and simultaneously for 2 or more seconds until K appears. Press the button the type of probe cycles through K (the default), J, T, R, S, and E types. The current mode is displayed on , an "S" will appear in the center of the left side of the LCD. press the screen. 5 Using the Pushbuttons The display defaults to the mode last used. For your convenience the meter defaults to the settings used during the last operation. REC HOLD REL ˚C ˚F T1 T2 6 Use the pushbuttons to control operation of the TMD90. The table assumes the TMD90 has been powered on with two thermocouples installed and is set to display (default) T1 on the primary display, type K thermocouple, and Record off. DA Button F B C E Description Turns the meter on and off. Press and release HOLD/REL and the Primary display (T1, T2, or T1-T2) freezes with HOLD displayed on top; Press for two or more seconds REL appears on top of LCD and the REL Primary display indicates the relative zero. Relative zero causes the value of the primary display to show as "000.0" , then only the amount of temperature change will be indicated. Relative temperatures can be recorded. Press HOLD/REL again and the unit returns to default. Press momentarily and the Primary display changes to T2 (Secondary screen displays T1); press momentarily again and it displays T1-T2; Secondary display alternates between T1 and T2; Press momentarily again and the instrument returns to default. Press momentarily and the backlight illuminates for approximately 30 seconds then turns off. Press momentarily and the unit toggles between Fahrenheit (the default) and Celsius temperatures; The current mode is indicated on the right side of the LCD. Press momentarily and the Relative Clock starts in the lower right screen. All other button functions are locked out except Power and Backlight. T1, T2, or T1-T2 is displayed on the Primary screen; The Secondary screen continues to update. Press momentarily again and the unit cycles through MAX and MIN (Maximum and Minimum recorded temperatures) and back to current temperature; The record mode is displayed on the LCD. Press and hold for three seconds to turn off the record function. RS-232 Output With optional software and cable, P/N TM-SW, the TMD90 can output measurement results to a computer with operating systems of Window 95/98/NT/2000/XP/ME. Features of the software: • Record up to 16,000 sample readings with real-time clock memory. • Programmable sample interval from 1 sec to 3,600 sec. • User defined maximum and minimum alarm settings. • Allow further analysis by downloading data to computer program. • All data is saved in .txt format for easy transformation into another software program. 7 Troubleshooting Problem Thermometer does not turn on. Solution • Check battery voltage and replace discharged battery. • Verify that battery clips grip the battery post tightly. • Insert missing thermocouple. Dashes appear in the T1 and T2 data screens. Dashes appear in T1 and/or T2 data screens with thermocouples inserted. • Measure resistance of thermocouples to ensure they are not broken internally. • Clean the thermocouple and restart. Temperature drifts from a known value in a controlled environment • Verify that thermocouple type matches the displayed icon. • Clean and dry the thermocouple blades and allow to air dry. • Confirm temperature with a known good thermocouple. Replace the defective thermocouple if required. 8 E Problem Relative clock does not start when you push the button. Dashes appear during a review of the maximum recorded value. Solution • Verify that the thermocouple is inserted correctly. • Thermocouple is not acknowledged. Check for intermittent or momentary thermocouple removal. • Check for HOLD icon on the display. Press button firmly. • Press and simultaneously before power on to disable the Auto power off feature. A A D A Data continues to update after you press the button. Instrument turns off while recording. Replacing the Battery Replace the 9 V battery when: • BAT appears on the right side of the display. • The meter will not turn on. • BAT appears when you turn on the backlight. Even if the battery was recently replaced, check the voltage level if there is no response from the meter. To replace the battery 1. Remove all thermocouples from the top of the meter. 2. Lay the instrument face down on a clean, flat surface. 3. Loosen the screw and remove the battery door. 4. Replace the battery. 5. Replace the battery door and tighten the screw. Remove the battery if you do not plan to use the meter for a month or more. Do not leave the battery in a meter that may be exposed to temperature extremes. 9 V ac Adapter For long term measurement recording, a 9 V AC adapter can be used in place of the battery. This can be purchased at any electronics store using these specs: Miniature power plug with inner conductor positive and outer conductor negative, 500 mA current rating. The 9 V battery can be left in the TMD90 when using the AC adapter. 9 Repair All test tools returned for warranty or non-warranty repair or for calibration should be accompanied by the following: your name, company’s name, address, telephone number, and proof of purchase. Additionally, please include a brief description of the problem or the service requested and include the test leads with the thermometer. Non-warranty repair or replacement charges should be remitted in the form of a check, a money order, credit card with expiration date, or a purchase order made payable to Meterman Test Tools. In-Warranty Repairs and Replacement – All Countries Please read the warranty statement that follows, and check your batteries and fuses before requesting repair. During the warranty period any defective test tool can be returned to your Meterman Test Tools distributor for an exchange for the same or like product. Please check the “Where to Buy” section on www.metermantesttools.com for a list of distributors near you. Additionally, in the United States and Canada In-Warranty repair and replacement units can also be sent to a Meterman Test Tools Service Center (see below for address). Non-Warranty Repairs and Replacement – US and Canada Non-warranty repairs in the United States and Canada should be sent to a Meterman Test Tools Service Center. Call Meterman Test Tools or inquire at your point of purchase for current repair and replacement rates. In USA In Canada Meterman Test Tools Meterman Test Tools 1420 75th Street SW 400 Britannia Rd. E. Unit #1 Everett, WA 98203 Mississauga, ON L4Z 1X9 Tel: 888-993-5853 Tel: 905-890-7600 Fax: 425-446-6390 Fax: 905-890-6866 Non-Warranty Repairs and Replacement – Europe European non-warranty units can be replaced by your Meterman Test Tools distributor for a nominal charge. Please check the “Where to Buy” section on www.metermantesttools.com for a list of distributors near you. European Correspondence Address* Meterman Test Tools Europe P.O. Box 1186 5602 BD Eindhoven The Netherlands *(Correspondence only – no repair or replacement available from this address. European customers please contact your distributor.) 10 WARRANTY The TMD90 Dual Input Thermometer is warranted against any defects of material or workmanship within a period of one (1) year following the date of purchase of the thermometer by the original purchaser or original user. Any thermometer claimed to be defective during the warranty period should be returned with proof of purchase to an authorized Meterman Test Tools Service Center or to the local Meterman Test Tools dealer or distributor where your thermometer was purchased. See Repair section for details. Any implied warranties arising out of the sale of a Meterman Test Tools thermometer, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited in duration to the above stated one (1) year period. Meterman Test Tools shall not be liable for loss of use of the thermometer or other incidental or consequential damages, expenses, or economical loss or for any claim or claims for such damage, expenses or economical loss. Some states do not allow limitations on how long implied warranties last or the exclusion or limitation of incidental or consequential damages, so the above limitations or exclusions may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state. Thermocouple Definitions Type J K R S T E Alloy Iron Fe Nickel-Chromium (Chromel) Ni-Cr Platinum-13% Rhodium Pt-13% Rh Platinum-10% Rhodium Pt-10% Rh Copper Cu Nickel-Chromium (Chromel) Ni-Cr 11 Alloy Copper-Nickel (Constantan) Cu-Ni Nickel-Aluminum Ni-AL Platinum Pt Platinum Pt Copper-Nickel (Constantan) Cu-Ni Copper-Nickel (Constantan) Cu-Ni Specifications T/C Type Range K -200 to 650 °C 651 to 1370 °C -328 to 1000 °F 1001 to 2498 °F -200 to 500 °C 501 to 760 °C -328 to 940 °F 941 to 1400 °F -200 to 390 °C -328 to 730 °F 0 to 1000 °C 1001 to 1760 °C 32 to 1000 °F 1001 to 3200 °F -200 to 380 °C 381 to 736 °C -328 to 720 °F 721 to 1832 °F J T R/S E Resolution 0.1 °C 1.0 °C 0.1 °F 1.0 °F 0.1 °C 1.0 °C 0.1 °C 1.0 °F 0.1 °C 0.1 °F 0.1 °C 1.0 °C 0.1 °F 1.0 °F 0.1 °C 1.0 °C 0.1 °F 1.0°F Accuracy (0.1% rdg + 0.7 °C) (0.1% rdg + 1.4 °F) (0.1% rdg + 0.7 °C) (0.1% rdg + 1.4 °F (0.1% rdg + 0.7 °C) (0.1% rdg + 1.4 °F) (0.3% rdg + 0.7 °C) (0.1% rdg + 1.4 °F) (0.3% rdg + 0.7 °C) (0.1% rdg + 1.4 °F) Additional Specifications OPERATING CONDITIONS 0 to 50 °C (32 to 122 °F) at 0 to 85 % relative humidity (non-condensing) Ambient Coefficient 0 to 18 °C and 28 to 50 °C (Ambient temperatures) For each °C ambient below 18 °C or above 28 °C, add the following tolerance into the accuracy spec: 0.01% of reading +0.03 °C (0.01% of reading +0.06 °F). 12 Thermomètre à double entrée TMD90 1 Français Consignes de sécurité ..........................................................................2 Symboles utilisés dans ce mode d'emploi..........................................2 Introduction.........................................................................................3 Affichage et commandes ......................................................................4 Fonctions.............................................................................................5 Arrêt automatique (mode de veille)....................................................5 Changer le type de thermocouple.......................................................5 Utilisation des boutons poussoirs......................................................6 Sortie RS-232 ...................................................................................8 Dépannage...........................................................................................9 Remplacement de la pile ....................................................................11 Adaptateur secteur de 9 V ..................................................................11 Réparation .........................................................................................12 GARANTIE .........................................................................................13 Définitions des thermocouples ...........................................................13 Spécifications ....................................................................................14 Autres caractéristiques.......................................................................14 Français Table des matières Consignes de sécurité • • • • • • Placer UNIQUEMENT des thermocouples dans l’entrée de thermocouple. S’assurer que l’appareil est configuré pour le type de thermocouple concerné. Vérifier que le thermocouple utilisé peut supporter les températures extrêmes auxquelles il sera exposé. Entretenir l’appareil avec soin et l’étalonner régulièrement Utilisez uniquement le thermomètre, en respectant les indications de ce manuel afin de ne pas entraver sa protection intégrée Remplacer la pile dès l’apparition de l'indicateur de pile faible apparaît pour éviter les mesures erronées qui posent des risques d'électrocution et de blessures. Avertissement Le thermocouple fourni n’est pas conçu pour être en contact avec des liquides ou des circuits électriques sous tension. XW Symboles utilisés dans ce mode d'emploi X T P Tension dangereuse W Double isolation J Conforme aux directives de l’UE 2 Se reporter au mode d’emploi Prise de terre Introduction Le thermomètre à entrée double modèle TMD90 accepte les thermocouples de types K, J, T, R, S et E. Il est doté d’un écran à triple affichage pour visualiser les résultats de mesure et l’horloge en temps relatif. Le logiciel et le câble optionnel RS-232 permettent de recueillir les données en temps réel vers un PC afin de les analyser ou de créer ultérieurement des rapports. Le TMD90 affiche tous les segments LCD (affichage à cristaux liquides) pendant environ 3 secondes lors du démarrage avant de fournir les informations sur tous les thermocouples connectés. Diverses combinaisons d’affichage sont possibles. L’écran LCD est divisé en trois sections distinctes ; une fenêtre principale (affichage primaire) en haut et deux fenêtres plus petites en bas (affichage secondaire et horloge relative). Les trois zones d’affichage sont actualisées en permanence avec les mesures de température et les informations de temps relatif. Plusieurs options permettent de configurer la présentation et le type des informations apparaissant sur l’écran LCD. • Les relevés de température basculent facilement entre les unités Fahrenheit et Celsius. • Un rétroéclairage illumine l’écran LCD pour les lectures en zones faiblement éclairées. gèle l’affichage de la fenêtre supérieure des • Le bouton données tout en permettant la mise à jour des informations dans les fenêtres du bas. • Un indicateur de pile faible apparaît également le cas échéant. • L’appareil adopte par défaut le dernier mode qui était sélectionné à la mise hors tension. Remarque : Si aucun thermocouple n’était connecté, quatre tirets (----) apparaissent sur l’affichage des températures. Contenu du coffret Thermomètre TMD90 1 Sonde de thermocouple de type K 1 Mode d’emploi 1 Boîtier de transport en plastique rigide 1 Pile de 9 V (installée) 1 A 3 Affichage et commandes 10 (T2) 10 (T1) 2 3 4 1 5 7 6 8 11 9 6 12 13 A B C D E F G H I J K REC L’horloge relative est active pour enregistrer les incidences de température min et max ou pour l’horodatage. MAX/MIN Les mesures de température minimum ou maximum actuellement contrôlées. HOLD/REL Gèle l’affichage primaire ou établit un zéro relatif pour les informations qu’elle contient. BAT Témoin de pile faible. °C °F Affiche la température en degrés Celsius ou Fahrenheit. T1-T2 Bascule entre l’affichage des informations T1 (primaire)/ T 2 (secondaire) et T2 (primaire) / T1 (secondaire), puis entre T 1-T 2(différentiel) sur l’affichage principal et les températures alternées T1/T2 sur la fenêtre secondaire. Affichage des données primaires. Affiche T1, T2 ou T1-T2( différentiel de température -TD ) ou un zéro relatif de T1, T2 ou TD. KJTRSE Sélectionne la référence d’entrée appropriée pour le thermocou ple utilisé. T1 et T2 doivent avoir le même type de thermocouple. Affichage des données secondaires. Affiche les températures de T2,T1 ou de T1 et T2. Entrée de thermocouple. Fiches à polarités positive et négative pour les sondes de thermocouple utilisées (type de lame) ; T1 à gauche ,T2 à droite. Affichage d’horloge relative. Affiche l’heure en heures, minutes et secondes (HHMMSS) à la pression de la touche REC, et l’heure relative de l’enregistrement des résultats MIN ou MAX. 4 L M Port de sortie RS-232. Câble et logiciel optionnel – Réf.TMSW. Entrée du convertisseur de courant alternatif à courant continu. W Attention Lire toutes les consignes de sécurité avant d’utiliser ce thermomètre. Fonctions Arrêt automatique (mode de veille) D A Le TMD90 s’arrête automatiquement environ 20 minutes après sa mise sous tension. Pour effectuer des enregistrements ou utiliser l’appareil pendant de longues périodes, désactivez le mode de veille en appuyant sur et sur simultanément avec l’appareil sous tension. A l’apparition de « n » au centre de l’affichage, relâchez la touche On. L’arrêt automatique est désactivé à la mise hors tension de l’appareil. C E D E Changer le type de thermocouple Appuyez sur le bouton et sur simultanément pendant au moins 2 secondes jusqu’à l’apparition de K. Appuyez sur le bouton et faites défiler les types de sondes, de K (par défaut) à J, T, R, S, et E. Le mode actuel s’affiche dans le coin gauche de l’écran LCD. , un « S » apparaît au centre de l’écran. Appuyez sur le bouton 5 Utilisation des boutons poussoirs L’affichage adopte par défaut le dernier mode utilisé. Pour plus de commodité, l’appareil utilise les paramètres utilisés pendant la dernière opération. REC HOLD REL ˚C ˚F T 1 T2 Utilisez les boutons poussoirs pour contrôler le fonctionnement du TMD 90. Dans le tableau suivant, on suppose que le TMD90 a été mis sous tension avec deux thermocouples installés et qu’il est réglé pour afficher (par défaut) T1 sur l’affichage primaire, le thermocouple de type K et l’enregistrement inactif. 6 DA Bouton F B C E Description Active ou désactive l’appareil. Appuyez sur HOLD/REL, puis relâchez et l'affichage principal (T1, T2 ou T1-T2) se fige avec HOLD affiché en haut. Maintenez appuyer pendant au moins deux secondes et REL apparaît en haut de l'écran à cristaux liquides et l'affichage principal REL indique le zéro relatif. Le zéro relatif oblige l'affichage principal à afficher la valeur « 000.0 », ensuite, seul le changement de température sera indiqué. Les températures relatives peuvent être enregistrées. Appuyez à nouveau sur HOLD/REL et l'appareil reprend sa valeur par défaut. Une pression momentanée et l’affichage primaire affiche T2 (la fenêtre secondaire affiche T1) ; une nouvelle pression momentanée permet d’afficher T1-T2 ; la fenêtre secondaire bascule entre T1 et T2 ; une dernière pression momentanée et l’instrument revient au réglage par défaut. Une pression momentanée et le rétroéclairage s’allume pendant une trentaine de secondes avant de s’éteindre. Une pression momentanée et l’appareil bascule entre les températures en Fahrenheit (par défaut) et Celsius ; le mode courant est indiqué à droite sur l’écran LCD. Une pression momentanée et l’horloge relative (REC) démarre dans la fenêtre inférieure droite, s’affiche dans le coin supérieur gauche de l’écran LCD – Toutes les autres fonctions sont verrouillées en dehors de l’alimentation et du rétroéclairage. T1, T2 ou T1-T2 s’affiche sur l’affichage primaire ; la fenêtre secondaire continue d’être actualisée. Une nouvelle pression momentanée et l’appareil répète les valeurs MAX et MIN (températures maximum et minimum enregistrées) et revient à la température courante ; le mode d’enregistrement apparaît sur l’écran LCD. Maintenez la touche enfoncée pendant trois secondes pour désactiver la fonction d’enregistrement. 7 Sortie RS-232 L’utilisation d’un câble et d’un logiciel optionnel, réf. T90/10-SW, permet au TMD90 de transmettre les résultats des mesures vers un ordinateur fonctionnant sous Windows 95/98/NT/2000/XP/ME. Fonctionnalités du logiciel : • Enregistre jusqu’à 16.000 valeurs exemples en mémoire avec horloge en temps réel. • Intervalle d’échantillonnage programmable de 1 s à 3600 s. • Paramètres d’alarme maximum et minimum définis par l’utilisateur. • Permet une analyse approfondie en téléchargeant les données vers un programme informatique. • Toutes les données sont enregistrées en format .txt permettant de les convertir facilement dans un autre programme d’application. 8 Dépannage Problème Le thermomètre ne s’active pas. Solution • Vérifiez la tension de la pile et remplacez la pile si elle est déchargée. • Assurez-vous que les pinces enserrent étroitement la borne de la pile. • Insérez le thermocouple manquant. Des tirets apparaissent dans le fenêtres de résultats T1 et T2. • Des tirets apparaissent dans les fenêtres de résultats T1 et/ou T2 avec les thermocouples insérés. • 9 Mesurez la résistance des thermocouples pour vérifier s’ils ne présentent pas de cassure interne. Nettoyez le thermocouple et recommencez. Problème La température dérive d’une valeur connue en environnement contrôlé E L’horloge relative ne démarre pas à la pression du bouton . Des tirets apparaissent pendant l’examen de la valeur maximale enregistrée. A Solution Vérifiez si le type de thermocouple correspond à l’icône affichée. • Nettoyez et séchez les lames de thermocouple et laissez sécher à l’air libre. • Confirmez la température avec un thermocouple identifié correct. Remplacez le thermocouple défectueux s’il y lieu. Vérifiez si le thermocouple est inséré correctement. • • Le thermocouple n'est pas reconnu. Vérifiez l’absence de retraits intermittents ou momentanés du thermocouple. Les données continuent d’être actualisées après la pression du bouton . • Recherchez l'icône HOLD sur l'affichage. Appuyez fermement sur le bouton . L’appareil s’éteint pendant l’enregistrement. • Appuyez sur et sur simultanément avant le démarrage pour désactiver la fonction de mise en veille automatique. A A D 10 Remplacement de la pile Remplacez la pile de 9 V lorsque : • BAT apparaît à droite sur l’affichage. • L’appareil ne s’allume pas. • BAT apparaît au démarrage du rétroéclairage. Même si la pile vient d’être remplacée, vérifiez son niveau de tension si l’appareil ne réagit pas. Pour remplacer la pile 1. Retirez tous les thermocouples du sommet de l’appareil. 2. Posez l’appareil face vers le bas sur une surface propre et plane. 3. Desserrez la vis et enlevez le capot de pile. 4. Remplacez la pile. 5. Replacez le capot de pile et serrez la vis. Retirez la pile si vous n’envisagez pas d’utiliser l’appareil pendant plus d’un mois. Ne laissez pas la pile dans l’appareil s’il risque d’être exposé à des températures extrêmes. Adaptateur secteur de 9 V Un adaptateur secteur de 9 V peut être utilisé à la place de la pile pour enregistrer les mesures à long terme. Cet adaptateur vendu dans les magasins d’électronique a les caractéristiques suivantes : un connecteur d’alimentation miniature doté d’un conducteur interne positif et d’un conducteur externe négatif, une intensité de 500mA. La pile de 9V peut être laissée dans le TMD90 pendant l’utilisation de l’adaptateur secteur. 11 Réparation Tous les outils de test renvoyés pour un étalonnage ou une réparation couverte ou non par la garantie doivent être accompagnés des éléments suivants : nom, raison sociale, adresse, numéro de téléphone et justificatif d'achat. Ajoutez également une brève description du problème ou du service demandé et incluez les cordons de test avec le thermomètre. Les frais de remplacement ou de réparation hors garantie doivent être acquittés par chèque, mandat, carte de crédit avec date d'expiration, ou par bon de commande payable à l'ordre de Meterman Test Tools. Remplacements et réparations sous garantie – Tous pays Veuillez lire la déclaration de garantie qui suit, et vérifiez les piles, batteries et fusibles avant de demander une réparation. Pendant la période de garantie, tout outil de test défectueux peut être renvoyé auprès de votre distributeur Meterman Test Tools pour être échangé contre un produit identique ou similaire. Consultez la section « Where to Buy » sur le site www.metermantesttools.com pour obtenir la liste des distributeurs dans votre région. Au Canada et aux Etats-Unis, les appareils devant être remplacé ou réparé sous garantie peuvent également être envoyés dans un centre de services (voir les adresses ci-dessous). Remplacements et réparations hors garantie – Canada et Etats-Unis Les appareils à réparer hors garantie au Canada et aux Etats-Unis doivent être envoyés dans un centre de services Meterman Test Tools. Appelez Meterman Test Tools ou renseignez-vous auprès de votre lieu d’achat pour connaître les tarifs en vigueur de remplacement ou de réparation. Aux Etats-Unis Au Canada Meterman Test Tools Meterman Test Tools 1420 75th Street SW 400 Britannia Rd. E. Unit #1 Everett, WA 98203 Mississauga, Ontario L4Z 1X9 Tél. : 888-993-5853 Tél. : 905-890-7600 Fax : 425-446-6390 Fax : 905-890-6866 Remplacements et réparations hors garantie – Europe Les appareils européens non couverts par la garantie peuvent être remplacés par votre distributeur Meterman Test Tools pour une somme nominale. Consultez la section « Where to Buy » sur le site www.metermantesttools.com pour obtenir la liste des distributeurs dans votre région. Adresse postale européenne* Meterman Test Tools Europe P.O. Box 1186 5602 B.D. Eindhoven Pays-Bas *(Réservée à la correspondance – Aucune réparation ou remplacement n’est possible à cette adresse. Nos clients européens doivent contacter leur distributeur.) 12 GARANTIE Le thermomètre à double entrée TMD90 est garanti contre tout défaut de fabrication ou de main d’œuvre pendant une période d’un (1) an à compter de la date d’achat du thermomètre par l’acheteur initial ou l’utilisateur initial. Tout thermomètre faisant l'objet d'un défaut pendant la période de garantie doit être renvoyé accompagné d’un justificatif d’achat auprès d’un centre de services agréé par Meterman Test ou du distributeur ou du revendeur local de Meterman Test Tools où l'achat du thermomètre a été effectué. Voir la section Réparation pour tous les détails. Toutes les garanties implicites résultant de la vente d’un thermomètre Meterman Test Tools, y compris mais sans s’y limiter les garanties de commercialisation ou d’adaptation à un usage particulier, sont limitées à la durée d’un (1) an déjà citée. Meterman Test Tools ne sera pas tenu responsable de la privation de jouissance du thermomètre ou d’autres dommages directs ou indirects, frais ou pertes économiques ni des poursuites engagées pour de tels dommages, frais ou pertes économiques. Certains pays n’admettent pas les limitations sur la durée des garanties implicites, ni sur l’exclusion ou la limitation des dommages directs ou indirects ; il est donc possible que les limitations ou exclusions de cette garantie ne s’appliquent pas dans votre cas. La présente garantie confère certains droits juridiques : la législation du pays ou de l’état peut vous en accorder d’autres. Définitions des thermocouples Type J Alliage Fer Fe K Nickel-Chrome (Chromel) Ni-Cr Platinum-13 % Rhodium Pt-13% Rh Platinum-10 % Rhodium Pt-10% Rh Cuivre Cu R B T E Alliage Cupro-Nickel (Constantan) Cu-Ni Nickel-Aluminium Ni-AL Platinum Pt Platinum Pt Cupro-Nickel (Constantan) Cu-Ni Cupro-Nickel (Constantan) Cu-Ni Nickel-Chrome (Chromel) Ni-Cr 13 Spécifications Type T/C Gamme Résolution Précision K -200 à 650 °C 651 à 1370 °C -328 à + 1000 °F 1001 à 2498 °F -200 à 500 °C 501 à 760 °C -328 à 940 °F 941 à 1400 °F -200 à 390 °C -328 à 730 °F 0 à 1000 °C 1001 à 1760 °C 32 à 1000 °F 1001 à 3200 °F -200 à 380 °C 381 à 736 °C -328 à 720 °F 721 à 1832 °F 0.1 °C 1.0 °C 0,1 °F 1,0 °F 0,1 °C 1,0 °C 0,1 °C 1,0 °F 0,1 °C 0,1 °F 0,1 °C 1,0 °C 0.1 °F 1.0 °F 0.1 °C 1.0 °C 0.1 °F 1.0°F (0,1 % de lecture + 0,7 °C) J T R/S E (0,1 % de lecture +1,4 °F) (0,1 % de lecture + 0,7 °C) (0,1 % de lecture +1,4 °F) (0,1 % de lecture + 0,7 °C) (0,1 % de lecture +1,4 °F) (0,3% de lecture + 0,7 °C) (0,1 % de lecture +1,4 °F) (0,3% de lecture + 0,7 °C) (0,1 % de lecture +1,4 °F) Autres caractéristiques CONDITIONS D’EXPLOITATION 0 à 50 °C (32 à 122 °F) avec 0 à 85 % d’humidité relative (sans condensation) Coefficient ambiant de 0 à 18 °C et de 28 à 50 °C (températures ambiantes) Pour chaque °C ambiant inférieur à 18 °C ou supérieur à 28 °C, ajouter la tolérance suivante à la caractéristique de précision : 0,01 % de lecture + 0,03 °C (0,01 % de lecture + 0,06 °F) 14 TMD90 Dual Input Thermometer 1 Deutsch Sicherheitsinformationen .....................................................................2 Symbole in diesem Handbuch ...........................................................2 Einleitung.............................................................................................3 Anzeige und Bedienelemente ................................................................4 Funktionen...........................................................................................5 Automatische Abschaltung (Ruhemodus) ..........................................5 Ändern des Thermoelementtyps ........................................................5 Verwenden der Tasten.......................................................................6 RS-232-Ausgang ..............................................................................8 Fehlerbehebung ...................................................................................9 Ersetzen der Batterie ..........................................................................11 9-V-Wechselstromadapter..................................................................11 Reparatur ..........................................................................................12 GARANTIE .........................................................................................13 Thermoelementdefinitionen ................................................................13 Spezifikationen...................................................................................14 Weitere Spezifikationen ......................................................................14 Deutsch Inhalt Sicherheitsinformationen • • • • • • NUR Thermoelemente in den Thermoelementeingang einführen. Sicherstellen, dass das Messgerät für den verwendeten Thermoelementtyp konfiguriert ist. Prüfen, ob das verwendete Thermoelement die Temperaturextreme, denen es ausgesetzt wird, aushalten kann. Das Messgerät ordnungsgemäß unterhalten und regelmäßig kalibrieren. Das Thermometer ausschließlich wie in diesem Handbuch beschrieben einsetzen, da sonst die im Thermometer integrierten Schutzeinrichtungen beeinträchtigt werden können. Zur Vermeidung falscher Messwerte, die zu Stromschlag oder Verletzungen führen können, die Batterie ersetzen, sobald die BAT eingeblendet wird. Warnung Das gelieferte Thermoelement darf nicht mit Flüssigkeiten oder stromführenden Stromkreisen in Kontakt kommen. XW Symbole in diesem Handbuch X T P Gefährliche Spannung Schutzisoliert W J Übereinstimmung mit EU-Richtlinien 2 Im Handbuch nachlesen. Erde, Masse Einleitung Das Modell TMD90 Dual Input Thermometer akzeptiert die Thermoelementtypen K, J, T, R, S und E. Das Gerät verfügt über einen Dreifach-Anzeigebildschirm zur Anzeige der Messergebnisse und des relativen Zeitgebers. Wahlfreie RS-232-Software und Kabel ermöglichen das Aufzeichnen von Echtzeitdaten auf einem PC für spätere Analysen und Berichte. Das TMD90 zeigt nach dem Einschalten alle LCD-Segmente (LCD = liquid crystal display) ungefähr 3 Sekunden lang an, und liefert dann Informationen zu beliebigen angeschlossenen Thermoelementen. Es gibt mehrere Anzeigekombinationen. Die LCD ist in drei gesonderte Abschnitte unterteilt: eine große obere Anzeige (Primäranzeige) und zwei kleinere untere Anzeigen (Sekundäranzeige und Anzeige für den relativen Zeitgeber). Die drei Anzeigebereiche werden kontinuierlich mit Informationen zu Temperaturmessungen und zum relativen Zeitgeber aktualisiert. Es gibt mehrere Optionen bezüglich Anzeige und Auswahl der auf der LCD dargestellten Informationen. • Temperaturmesswerte können einfach zwischen Fahrenheit und Celsius umgeschaltet werden. • Eine Hintergrundbeleuchtung beleuchtet die LCD für Ablesungen in Arbeitsbereichen mit wenig Licht. Taste friert die oberen Anzeigedaten ein und ermöglicht • Die fortgesetzte Aktualisierung der Informationen der unteren Anzeigen. • Darüber hinaus wird ggf. auch eine Anzeige für schwache Batterie angezeigt. • Das Messgerät kehrt beim Einschalten zu dem Modus zurück, der vor dem Ausschalten zuletzt verwendet wurde. Hinweis: Wenn keine Thermoelemente angeschlossen sind, werden in der Temperaturanzeige vier Gedankenstriche (----) angezeigt. Lieferumfang TMD90 Thermometer 1 Typ K Thermoelementsonde 1 Bedienungsanleitung 1 Transportbehälter 1 9 V Batterie (installiert) 1 A 3 Anzeige und Bedienelemente 10 (T2) 10 (T1) 2 3 4 1 5 7 6 8 11 9 6 12 13 A B C D E F G H I J K REC Relativer Zeitgeber aktiviert zur Aufzeichnung der Min.- und Max.-Temperatur und für Zeitstempelvergabe. MAX/MIN Min.- bzw. Max.-Temperaturmesswerte werden überwacht. HOLD/REL Friert Primäranzeige ein bzw. setzt einen relativen Nullpunkt für die Daten in der Primäranzeige. BAT Anzeige für schwache Batterie. °C °F Zeigt Temperatur in Grad Celsius oder Grad Fahrenheit an. T1-T2 Schaltet Anzeigeinformationen um: von T1 (Primäranzeige)/ T2 (Sekundäranzeige) zu T2 (Primäranzeige) / T1 (Sekundäranzeige), dann zu T1-T2 (Differenzial) auf der Primäranzeige und wechselweise T1/T2-Temperaturen auf der Sekundäranzeige. Primärdatenanzeige. Zeigt T1, T2 oder T1-T2 (Temperaturdifferenzial -TD ) oder einen relativen Nullpunkt von T1, T2 oder TD an. KJTRSE Wählt korrekte Eingangsreferenz für das verwendete Thermoelement aus. T1 und T2 müssen denselben Thermoelementtyp aufweisen. Sekundärdatenanzeige. Zeigt T2, T1 oder T1 und T2 Temperaturen an. Thermoelementeingang. Positiv und negativ polarisierte Stecker für die verwendeten Thermoelementsonden (Spatenstecker); T1 links, T2 rechts. Anzeige des relativen Zeitgebers. Zeigt die Uhrzeit in Stunden, Minuten und Sekunden (HHMMSS) an, wenn REC gedrückt wird, und die relative Zeit der Aufzeichnung der MIN- bzw. MAX-Daten an. 4 L M RS-232-Ausgangsanschluss. Wahlfreie Software und Kabel – Teile-Nr. TM-SW. Wechselstrom-/Gleichstromwandlereingang. W Vorsicht Vor Gebrauch dieses Thermometers alle Sicherheitsinformationen lesen. Funktionen Automatische Abschaltung (Ruhemodus) DA Das TMD90 schaltet ungefähr 20 Minuten nach dem Einschalten automatisch ab. Für Aufzeichnungen bzw. Betrieb über längere Zeitperioden kann der Ruhemodus durch gleichzeitiges Drücken der Tasten und während des Einschaltens deaktiviert werden. Wenn „n“ in der Mitte der Anzeige erscheint, kann die ON-Taste losgelassen werden. Die automatische Abschaltung wird deaktiviert, wenn das Messgerät ausgeschaltet wird. ED C Ändern des Thermoelementtyps Die Tasten und gleichzeitig mindestens 2 Sekunden lang drücken, bis K angezeigt wird. Die Taste drücken, sodass die möglichen Sondentypen durchlaufen werden: K (Standard), J, T, R, S und E. Der aktuelle Modus wird auf der linken Seite der LCD angezeigt. drücken, sodass ein „S“ in der Mitte der Anzeige erscheint. E 5 Verwenden der Tasten Die Anzeige kehrt standardmäßig zum zuletzt verwendeten Modus zurück. Das Messgerät verwendet standardmäßig die Einstellungen, die während des letzten Einsatzes verwendet wurden. REC HOLD REL ˚C ˚F T1 T2 6 Die Tasten zur Steuerung des TMD90 verwenden. Die Tabelle setzt voraus, dass das TMD90 über zwei installierte Thermoelemente verfügt, eingeschaltet ist, zur Anzeige von T1 (Standard) auf der Primäranzeige und Typ K Thermoelement eingestellt ist und „Record“ deaktiviert ist. DA Taste F B C E Beschreibung Schaltet das Messgerät ein und aus. Kurzzeitiges Drücken von HOLD/REL bewirkt, dass die Primäranzeige (T1, T2 bzw. T1-T2) einfriert, und HOLD wird oben angezeigt. Drücken für 2 oder mehr Sekunden bewirkt, dass REL oben in der LCD angezeigt wird, und die REL-Primäranzeige zeigt den relativen Nullpunkt an. Relativer Nullpunkt bewirkt, dass der Wert der Primäranzeige als "000,0" angezeigt wird, danach wird nur die Temperaturveränderung angezeigt. Relative Temperaturen können aufgezeichnet werden. Nochmaliges Drücken von HOLD/REL bewirkt, dass die Einheit auf Standard zurückschaltet. Kurzzeitiges Drücken bewirkt, dass die Primäranzeige auf T2 (Sekundäranzeige zeigt T1 an) schaltet; nochmaliges kurzzeitiges Drücken bewirkt, dass T1-T2 angezeigt wird; Sekundäranzeige wechselt zwischen T1 und T2; nochmaliges kurzzeitiges Drücken bewirkt, dass das Instrument auf Standard zurückschaltet. Kurzzeitiges Drücken bewirkt, dass die Hintergrundbeleuchtung ungefähr 30 Sekunden lang leuchtet und dann ausschaltet. Kurzzeitiges Drücken bewirkt, dass die Einheit zwischen Fahrenheit- (Standard) und CelsiusTemperaturen umschaltet; der aktuelle Modus rechts auf der LCD angezeigt. Kurzzeitiges Drücken bewirkt, dass der relative Zeitgeber unten rechts auf der Anzeige zu laufen beginnt; ... wird oben links auf der LCD angezeigt alle übrigen Tastenfunktionen mit Ausnahme von „Power“ und „Backlight“ sind ausgesperrt. T1, T2 bzw. T1-T2 wird in der Primäranzeige angezeigt; die Sekundäranzeige wird weiterhin aktualisiert. Nochmaliges kurzzeitiges Drücken bewirkt, dass die Einheit die MAX- und MIN-Werte (aufgezeichnete Maximal- und Minimaltemperaturen) durchläuft und zur aktuellen Temperaturanzeige zurückkehrt. Die Taste drücken und 3 Sekunden halten, um die Aufzeichnungsfunktion auszuschalten. 7 RS-232-Ausgang Mit wahlfreier Software und Kabel, Teile-Nr. T90/10-SW, kann das TMD90 Messergebnisse auf einen Computer unter Windows 95/98/NT/2000/XP/ME ausgeben. Zu den Merkmalen der Software gehören: • Aufzeichnung von bis zu 16000 Messwerten in Speicher mit Echtzeituhr. • Programmierbares Intervall: 1 bis 3600 Sekunden. • Benutzerdefinierte Maximal- und Minimal-Alarmeinstellungen. • Ermöglicht weitere Analysen durch Herunterladen von Daten auf einen Computer Verarbeitung mit einem Programm. • Alle Daten werden im .txt-Format gespeichert, sodass sie einfach in andere Softwareprogramme übernommen werden können. 8 Fehlerbehebung Problem Thermometer lässt sich nicht einschalten. Lösung • Batteriespannung prüfen und ggf. aufgebrauchte Batterie ersetzen. • Sicherstellen, dass Batterieklemmen die Batteriepole sicher festhalten. • Fehlendes Thermoelement einsetzen. Gedankenstriche erscheinen in den T1- und T2Datenanzeigen. • Widerstand der Thermoelemente messen, um interne Defekte auszuschließen. • Das Thermoelement reinigen und neu beginnen. Gedankenstriche erscheinen in den T1- und/oder T2Datenanzeigen und Thermoelemente sind eingesetzt. 9 Problem Temperatur schwankt in einer kontrollierten Umgebung von einem bekannten Wert. E Der relative Zeitgeber startet nicht, wenn die Taste gedrückt wird. Gedankenstriche erscheinen während der Anzeige des aufgezeichneten Maximalwerts. A Lösung • Prüfen, ob Thermoelementtyp mit dem angezeigten Symbol übereinstimmt. • Die Thermoelementblätter reinigen und trocknen und die Einheit lufttrocknen lassen. • Temperatur mit einem einwandfreien Thermoelement bestätigen. Das defekte Thermoelement nötigenfalls ersetzen. • Sicherstellen, dass das Thermoelement korrekt eingesetzt ist. • Thermoelement ist nicht bestätigt. Auf zeitweilige/kurzzeitige Trennung des Thermoelements prüfen. Daten werden nach Drücken der Taste weiterhin aktualisiert. • Prüfen, ob das HOLDSymbol HOLD auf der Anzeige erscheint. Die Taste fest drücken. Instrument schaltet sich während einer Aufzeichnung ab. • Die Tasten und vor dem Einschalten gleichzeitig drücken, um die Funktion „Automatische Abschaltung“ zu deaktivieren. A D A 10 Ersetzen der Batterie Die 9-V-Batterie ersetzen, wenn: • BAT rechts auf der Anzeige eingeblendet wird. • Das Messgerät nicht eingeschaltet werden kann. • BAT eingeblendet wird, wenn die Hintergrundbeleuchtung aktiviert wird. Selbst wenn die Batterie kürzlich ersetzt wurde, den Spannungspegel prüfen, falls das Messgerät nicht reagiert. Ersetzen der Batterien 1. Alle Thermoelemente vom Oberteil des Messgeräts entfernen. 2. Das Instrument mit der Vorderseite nach unten auf eine saubere, flache Oberfläche legen. 3. Die Schraube lösen und die Batteriefachabdeckung abnehmen. 4. Die Batterie ersetzen. 5. Die Batteriefachabdeckung wieder anbringen und die Schraube anziehen. Die Batterie entfernen, falls das Messgerät voraussichtlich einen Monat oder länger nicht verwendet wird. Die Batterie nicht in einem Messgerät belassen, das möglicherweise extremen Temperaturen ausgesetzt wird. 9-V-Wechselstromadapter Für Langzeit-Messwertaufzeichnung kann ein 9-VWechselstromadapter an Stelle der Batterie eingesetzt werden. Ein solcher Adapter kann unter Berücksichtigung der folgenden Spezifikationen in einem beliebigen Elektronikgeschäft gekauft werden: Mini-Netzstecker mit positivem Innenleiter, negativen Außenleiter und 500 mA Nennstromstärke. Die 9-V-Batterie kann im TMD90 belassen werden, wenn der Wechselstromadapter verwendet wird. 11 Reparatur Zu allen Geräten, die zur Reparatur oder Kalibrierung im Rahmen der Garantie oder außerhalb der Garantie eingesendet werden, muss folgendes beigelegt werden: Name des Kunden, Firmenname, Adresse, Telefonnummer und Kaufbeleg. Zusätzlich eine kurze Beschreibung des Problems oder der gewünschten Wartung sowie die Messleitungen dem Thermometer beilegen. Die Gebühren für Reparaturen außerhalb der Garantie oder für den Ersatz von Instrumenten müssen als Scheck, Geldanweisung, Kreditkarte (Kreditkartennummer mit Ablaufdatum) beglichen werden oder es muss ein Auftrag an Meterman Test Tools formuliert werden. Garantiereparaturen oder -austausch – alle Länder Bitte die nachfolgende Garantieerklärung lesen, und die Batterien und Sicherungen prüfen, bevor Reparaturen angefordert werden. Während der Garantieperiode können alle defekten Geräte zum Umtausch gegen dasselbe oder ein ähnliches Produkt an den Meterman Test Tools-Distributor gesendet werden. Ein Verzeichnis der zuständigen Distributoren ist im Abschnitt „Where to Buy“ (Verkaufsstellen) auf der Website www.metermantesttools.com zu finden. Darüber hinaus können in den USA und in Kanada Geräte an ein Meterman Test Tools Service-Center (Adresse siehe weiter unten) zur Reparatur oder zum Umtausch eingesendet werden. Reparaturen und Austausch außerhalb der Garantie – USA und Kanada Für Reparaturen außerhalb der Garantie in den Vereinigten Staaten und in Kanada werden die Geräte an ein Meterman Test Tools Service-Center gesendet. Auskunft über die derzeit geltenden Reparatur- und Austauschgebühren erhalten Sie von Meterman Test Tools oder der Verkaufsstelle. In den USA: In Kanada: Meterman Test Tools Meterman Test Tools 1420 75th Street SW 400 Britannia Rd. E. Unit #1 Everett, WA 98203 Mississauga, ON L4Z 1X9 Tel.: 888-993-5853 Tel.: 905-890-7600 Fax: 425-446-6390 Fax: 905-890-6866 Reparaturen und Austausch außerhalb der Garantie – Europa Geräte außerhalb der Garantie können durch den zuständigen Meterman Test Tools-Distributor gegen eine Gebühr ersetzt werden. Ein Verzeichnis der zuständigen Distributoren ist im Abschnitt „Where to Buy“ (Verkaufsstellen) auf der Website www.metermantesttools.com zu finden. Korrespondenzanschrift für Europa* Meterman Test Tools Europe P.O. Box 1186 5602 BD Eindhoven Niederlande *(Nur Korrespondenz – keine Reparaturen, kein Umtausch unter dieser Anschrift. Kunden in Europa wenden sich an den zuständigen Distributor.) 12 GARANTIE Es wird gewährleistet, dass das TMD90 Dual Input Thermometer innerhalb eines Zeitraums von einem (1) Jahr ab dem Datum des Kaufes des Thermometers durch den Erstkäufer oder Erstnutzer frei von Material- oder Fertigungsfehlern ist. Thermometer, die während der Garantieperiode als defekt angegeben werden, müssen mit dem Kaufbeleg an ein autorisiertes Meterman Test Tools Service-Center oder den örtlichen Meterman Test Tools-Händler/-Distributor (bei dem das Thermometer gekauft wurde) gesendet werden. Nähere Einzelheiten siehe Abschnitt „Reparatur“. Alle aus dem Kauf eines Meterman Test Tools-Thermometers abgeleiteten Garantien, einschließlich, aber nicht beschränkt auf die abgeleiteten Garantien der Marktfähigkeit oder Eignung für einen bestimmten Zweck, sind auf die Dauer des oben angegebenen Zeitraums von einem (1) Jahr beschränkt. Meterman Test Tools haftet nicht für Nutzungsausfall des Thermometers oder andere Begleit- oder Folgeschäden, Ausgaben oder wirtschaftliche Verluste oder für jegliche Ansprüche bezüglich solcher Schäden, Ausgaben oder wirtschaftlicher Verluste. In einigen Ländern ist die zeitliche Begrenzung der abgeleiteten Gewährleistung sowie der Ausschluss oder die Begrenzung von Begleit- oder Folgeschäden nicht zulässig, so dass die oben genannten Einschränkungen und Ausschlüsse nicht für jeden Käufer gelten. Diese Gewährleistung gibt dem Eigentümer bestimmte Rechte sowie möglicherweise andere Rechte, die von Land zu Land verschieden sind. Thermoelementdefinitionen Typ J Material Eisen Fe K Nickel-Chrom (Chromel) Ni-Cr Platin-13 % Rhodium Pt-13% Rh Platin-10 % Rhodium Pt-10% Rh Kupfer Cu R S T E Nickel-Chrom (Chromel) Ni-Cr 13 Material Kupfer-Nickel (Konstantan) Cu-Ni Nickel-Aluminium Ni-AL Platin Pt Platin Pt Kupfer-Nickel (Konstantan) Cu-Ni Kupfer-Nickel (Konstantan) Cu-Ni Spezifikationen TE-Typ Bereich K -200 bis 650 °C 651 bis 1370 °C -328 bis 1000 °F 1001 bis 2498 °F -200 bis 500 °C 501 bis 760 °C -328 bis 940 °F 941 bis 1400 °F -200 bis 390 °C -328 bis 730 °F 0 bis 1000 °C 1001 bis 1760 °C 32 bis 1000 °F 1001 bis 3200 °F -200 bis 380 °C 381 bis 736 °C -328 bis 720 °F 721 bis 1832 °F J T R/S E Auflösung 0,1 °C 1,0 °C 0,1 °F 1,0 °F 0,1 °C 1,0 °C 0,1 °C 1,0 °F 0,1 °C 0,1 °F 0,1 °C 1,0 °C 0,1 °F 1,0 °F 0,1 °C 1,0 °C 0,1 °F 1,0°F Genauigkeit (0,1% Anz. + 0,7 °C) (0,1% Anz. + 1,4 °F) (0,1% Anz. + 0,7 °C) (0,1% Anz. + 1,4 °F (0,1% Anz. + 0,7 °C) (0,1% Anz. + 1,4 °F) (0,3% Anz. + 0,7 °C) (0,1% Anz. + 1,4 °F) (0,3% Anz. + 0,7 °C) (0,1% Anz. + 1,4 °F) Weitere Spezifikationen BETRIEBSBEDINGUNGEN 0 bis 50 °C (32 bis 122 °F) bei 0 bis 85 % relativer Feuchtigkeit (nichtkondensierend). Umgebungskoeffizient 0 bis 18 °C und 28 bis 50 °C (Umgebungstemperatur). Pro °C Umgebungstemperatur unterhalb von 18 °C bzw. oberhalb von 28 °C die folgende Toleranz zur Genauigkeitsspezifikation hinzufügen: 0,01 % Ablesung +0,03 °C (0,01 % Ablesung +0,06 °F) 14 Termometro a due ingressi TMD90 1 Italiano Informazioni sulla sicurezza..................................................................2 Simboli adoperati nel presente manuale.............................................2 Introduzione ........................................................................................3 Display e comandi................................................................................4 Funzioni...............................................................................................5 Spegnimento automatico (modalità di risparmio energetico) ..............5 Cambio del tipo di termocoppia .........................................................5 Uso dei pulsanti ................................................................................6 Uscita RS-232...................................................................................8 Soluzione dei problemi.........................................................................9 Sostituzione della pila.........................................................................11 Adattatore di c.a. da 9 V .....................................................................11 Riparazioni.........................................................................................12 GARANZIA .........................................................................................13 Descrizione delle termocoppie ............................................................13 Dati tecnici......................................................................................14 Dati tecnici aggiuntivi .........................................................................14 Italiano Indice Informazioni sulla sicurezza • • • • • • All’ingresso per termocoppie collegare SOLO termocoppie. Accertarsi che il termometro sia configurato per l’uso della termocoppia a disposizione. Assicurarsi che la termocoppia in uso possa sostenere le temperature estreme alle quali sarà sottoposta. Mantenere in modo appropriato il termometro e calibrarlo regolarmente. Usare il termometro solo come specificato nel presente manuale, altrimenti si può compromettere la protezione offerta dallo strumento. Sostituire le pile non appena si visualizza il simbolo BAT, per evitare letture errate che potrebbero causare scosse elettriche e lesioni. Avvertenza La termocoppia fornita non deve entrare in contatto con liquidi o circuiti elettrici sotto tensione. XW Simboli adoperati nel presente manuale X T P Alta tensione Isolamento doppio W J Conforme alle direttive UE 2 Consultare il manuale Massa di terra Introduzione Il modello TMD90 è un termometro a due ingressi che consente il collegamento di termocoppie di tipo K, J, T, R, S ed E. Dispone di un display diviso in tre sezioni per la visualizzazione delle misure e dell’orologio relativo. Il software e il cavo RS-232 opzionali permettono la trasmissione di dati in tempo reale a un PC per l’ulteriore analisi e rapporti. All’accensione, il TMD90 visualizza tutti i segmenti del display a cristalli liquidi per circa 3 secondi e quindi fornisce informazioni sulle termocoppie eventualmente collegate. Sono disponibili numerose configurazioni di visualizzazione. Il display a cristalli liquidi è diviso in tre sezioni distinte: una più ampia nella parte superiore (visualizzazione principale) e due più piccole nella parte inferiore (visualizzazione secondaria e orologio relativo). Le tre aree di visualizzazione vengono aggiornate continuamente con le misure della temperatura e le informazioni di tempo relativo. Lo strumento offre diverse opzioni relative al tipo e al formato dei dati presenti sul display. • Le letture di temperatura possono essere facilmente convertite da gradi Fahrenheit a gradi centigradi e viceversa. • La retroilluminazione rende luminoso il display a cristalli liquidi per una facile lettura in ambienti scarsamente illuminati. blocca i dati della sezione superiore del display, • Il pulsante mentre la sezione inferiore continua ad aggiornarsi. • L’indicatore di pila scarica è anch’esso aggiornato secondo la necessità. • Il termometro attiva automaticamente l’ultima modalità selezionata prima dello spegnimento. Nota: se non vi sono termocoppie collegate, il visualizzatore della temperatura presenta quattro trattini (----). Contenuto della confezione Termometro TMD90 1 Sonda per termocoppia tipo K 1 Manuale d’uso 1 Custodia di plastica rigida per il trasporto 1 Pila da 9 V (installata) 1 A 3 Display e comandi 10 (T2) 10 (T1) 2 3 4 1 5 7 6 8 11 9 6 12 13 A B C D E F G H I J K L M REC L’orologio relativo è attivo per la registrazione di temperature minime e massime o per la marcatura temporale. MAX/MIN Permette il monitoraggio delle misure di temperatura massima o minima. HOLD/REL Blocca il visualizzatore principale o stabilisce lo zero relativo per i dati in esso presentati. BAT Indicatore di pila scarica. °C °F Visualizza la temperatura in gradi Fahrenheit o in gradi centigradi. T1-T2 Alterna la visualizzazione come segue: da T1 (principale)/T2 (secondario) a T2 (principale)/T1 (secondario), quindi a T1-T2 (differenziale) su visualizzatore principale e alternanza di temperature T1/T2 sul visualizzatore secondario. Visualizzatore principale. Mostra gli ingressi T1, T2 o T1-T2 (temperatura differenziale - TD) o lo zero relativo di T1, T2 o TD. KJTRSE Permette di selezionare il riferimento in ingresso corretto per la termocoppia in uso. T1 e T2 devono essere termocoppie dello stesso tipo. Visualizzatore secondario. Indica le temperature T2, T1 o T1 e T2. Ingressi delle termocoppie. Spinotti polarizzati positivi e negativi per le sonde delle termocoppie usate (tipo a lama): T1 a sinistra, T2 a destra. Visualizzatore dell’orologio relativo. Presenta le ore, i minuti e i secondi (HHMMSS) quando si preme REC e l’orario relativo in cui i dati MIN o MAX sono stati registrati. Uscita RS-232. Software e cavo opzionali (n. parte TM-SW). Ingresso del convertitore c.a./c.c. 4 W Attenzione Prima di usare il termometro, leggere la sezione “Informazioni sulla sicurezza”. Funzioni Spegnimento automatico (modalità di risparmio energetico) DA Il TMD90 si spegne automaticamente 20 minuti circa dopo l’accensione. Per registrazioni od operazioni che durano più a lungo, è possibile disattivare la modalità di risparmio energetico premendo contemporaneamente i pulsanti e con lo strumento acceso. Quando al centro dello schermo appare il simbolo “n”, si può rilasciare il pulsante On. La funzione di spegnimento automatico rimane disattivata quando si spegne il termometro. ED C Cambio del tipo di termocoppia Premere contemporaneamente i pulsanti e per almeno 2 secondi, finché non appare l’indicatore K. Premere il pulsante per visualizzare uno alla volta i tipi di sonda: K (il tipo predefinito), J, T, R, S ed E. La modalità corrente appare sul lato sinistro del display a cristalli liquidi. Premere : al centro dello schermo appare una “S”. E 5 Uso dei pulsanti Il display visualizza automaticamente l’ultima modalità selezionata. Inoltre, per comodità, il termometro richiama le impostazioni usate durante l’ultima operazione. REC HOLD REL ˚C ˚F T 1 T2 6 Usare i pulsanti per comandare le varie funzioni del termometro TMD90. Nella tabella si presume quanto segue: il TMD90 è stato acceso, sono state collegate due termocoppie, il visualizzatore principale presenta la termocoppia T1 (predefinita), la termocoppia è di tipo K e la registrazione è disattivata. Pulsante Descrizione DA F B C E Accende e spegne il termometro. Premendo brevemente il pulsante REL/HOLD, il visualizzatore principale (T1, T2 o T1-T2) si blocca e sopra di esso appare l’indicatore HOLD; premendolo per 2 secondi o più, nella parte superiore del display a cristalli liquidi appare l'indicatore REL e il visualizzatore principale (REL) presenta lo zero relativo. Lo zero relativo porta a "000.0" il valore nel visualizzatore principale; dopodiché, il display indica solo le variazioni di temperatura. In questa modalità è possibile registrare le temperature relative. Premere di nuovo il pulsante REL/HOLD per riportare lo strumento all'impostazione predefinita. Premendo brevemente il pulsante, il visualizzatore principale passa all’ingresso T2 (lo schermo secondario visualizza T1). Premendolo di nuovo brevemente, sul visualizzatore principale appare T1-T2 (quello secondario alterna tra T1 e T2). Premendolo una terza volta, lo strumento ritorna all’impostazione predefinita. Premendo brevemente il pulsante, la retroilluminazione si accende per circa 30 secondi e poi si spegne. Premendo brevemente il pulsante, il display da gradi Fahrenheit (unità predefinita) a gradi centigradi e viceversa. L’unità corrente è indicata sul lato destro del display. Premendo brevemente il pulsante, nell’angolo inferiore destro del display si avvia l’orologio relativo, mentre nell’angolo superiore sinistro si accende l’indicatore. Tutte le altre funzioni sono bloccate, ad eccezione dei pulsanti di alimentazione e di retroilluminazione. Il visualizzatore principale indica T1, T2 o T1-T2, mentre quello secondario continua ad aggiornarsi. Premendolo di nuovo brevemente, lo strumento passa dalla visualizzazione della temperatura massima registrata (MAX) a quella minima (MIN) e poi di nuovo a quella corrente. Il display visualizza la modalità di registrazione. Per disattivarla, tenere premuto il pulsante per tre secondi. 7 Uscita RS-232 Con il software e il cavo RS-232 opzionali (n. parte T90/10-SW), il TMD90 è in grado di inviare i risultati delle misure a un computer con sistema operativo Windows 95/98/NT/2000/XP/ME. Tra le caratteristiche del software, vi sono: • registrazione di un massimo di 16.000 letture campione con memoria dell’orologio di tempo reale; • intervallo di campionamento programmabile da 1 secondo a 3.600 secondi; • limiti di allarme minimo e massimo definiti dall’utente; • possibilità di analisi ulteriori mediante lo scaricamento di dati in un programma del computer; • memorizzazione di tutti i dati nel formato di testo (.txt) per agevolarne l’importazione in un altro programma software. 8 Soluzione dei problemi Problema Il termometro non si accende. Soluzione • Controllare la tensione della pila e sostituirla se è scarica. • Verificare che i terminali della pila facciano un buon contatto con i morsetti. I visualizzatori dei dati T1 e T2 presentano dei trattini. • Collegare la termocoppia mancante. I visualizzatori dei dati T1 e/o T2 presentano dei trattini benché le termocoppie siano collegate. • Misurare la resistenza delle termocoppie per accertarsi che non vi siano dei guasti interni. • Pulire le termocoppie e riprovare. 9 Problema Si verifica una deriva della temperatura da un valore noto in un ambiente controllato. E Soluzione • Verificare che il tipo di termocoppia corrisponda all’icona visualizzata. • Pulire e asciugare le lame delle termocoppie e lasciare che si asciughino all’aria. • Verificare la temperatura con una termocoppia funzionante. Sostituire la termocoppia eventualmente difettosa. L’orologio relativo non parte quando si preme il pulsante . • Verificare che la termocoppia si stata inserita bene. Appaiono dei trattini durante la revisione del valore massimo registrato. • La termocoppia non è stata riconosciuta. Verificare che non vi sia un contatto intermittente o che la termocoppia non sia stata rimossa temporaneamente. A I dati continuano ad aggiornarsi anche dopo aver premuto il pulsante . • Verificare la presenza dell'icona HOLD sul display. Premere il pulsante con decisione. Lo strumento si spegne durante la registrazione. • Premere contemporaneamente i pulsanti e prima di accendere lo strumento per disattivare la funzione di spegnimento automatico. A 10 DA Sostituzione della pila Sostituire la pila da 9 V quando: • l’icona BAT appare nell’angolo destro del display; • il termometro non si accende; • l’icona BAT appare quando si accende la retroilluminazione. Se il termometro non risponde ai comandi, controllare la tensione della pila, anche se è stata sostituita di recente. Sostituzione della pila 1. Scollegare tutte le termocoppie dalla parte superiore del termometro. 2. Appoggiare lo strumento su una superficie pulita e in piano, con la parte anteriore rivolta verso il basso. 3. Rimuovere lo sportello dello scomparto della pila, dopo averne allentato la vite. 4. Sostituire la batteria. 5. Rimontare lo sportello e serrare la vite. Se si prevede di non usare il termometro per un mese o più, rimuovere la pila. Rimuoverla anche nel caso in cui lo strumento venga esposto a temperature estreme. Adattatore di c.a. da 9 V Per la registrazione a lungo termine delle misure, anziché alimentare lo strumento a batteria si può usare un adattatore di c.a. a 9 V. Questo accessorio può essere acquistato in un qualsiasi negozio di componenti elettronici, ma deve soddisfare i seguenti requisiti: spinotto di alimentazione mini-plug con conduttore interno positivo e conduttore esterno negativo, 500 mA di corrente nominale. La pila da 9 V può essere lasciata nel TMD90 durante l’uso dell’adattatore. 11 Riparazioni A tutti gli strumenti di misura restituiti per interventi in garanzia o non coperti dalla garanzia, oppure per la taratura, devono essere allegate le seguenti informazioni: il proprio nome e quello dell’azienda, indirizzo, numero telefonico e scontrino. Allegare anche una breve descrizione del problema o dell’intervento richiesto e i cavetti. Gli importi dovuti per sostituzioni o riparazioni non coperte dalla garanzia vanno versati tramite assegno, vaglia bancario, carta di credito con data di scadenza od ordine di acquisto all’ordine di Meterman Test Tools. Sostituzioni e riparazioni in garanzia – Tutti i Paesi Si prega di leggere la seguente garanzia e di controllare la pila e i fusibili prima di richiedere una riparazione. Durante il periodo di garanzia, si può restituire uno strumento difettoso al rivenditore Meterman Test Tools per ricevere un prodotto identico o analogo. Nella sezione “Where to Buy” del sito c’è un elenco dei distributori più vicini. Negli Stati Uniti e nel Canada gli strumenti da sostituire o riparare in garanzia possono essere inviati anche a un centro di assistenza Meterman Test Tools (l’indirizzo è più avanti). Sostituzioni e riparazioni non coperte dalla garanzia – Usa e Canada Per riparazioni non coperte dalla garanzia, negli Stati Uniti e nel Canada lo strumento deve essere inviato a un centro di assistenza Meterman Test Tools. Rivolgersi alla Meterman Test Tools o al rivenditore per informazioni sui costi delle riparazioni e sostituzioni. USA Canada Meterman Test Tools Meterman Test Tools 1420 75th Street SW 400 Britannia Rd. E. Unit #1 Everett, WA 98203 Mississauga, ON L4Z 1X9 Tel: 888 993 5853 Tel: 905 890 7600 Fax: 425 446 6390 Fax: 905 890 6866 Sostituzioni e riparazioni non coperte dalla garanzia – Europa Gli strumenti acquistati in Europa e non coperti dalla garanzia possono essere sostituiti dal rivenditore Meterman Test Tools per un importo nominale. Nella sezione “Where to Buy” del sito c’è un elenco dei distributori più vicini. Recapito postale europeo* Meterman Test Tools Europe P.O. Box 1186 5602 BD Eindhoven Paesi Bassi *(Solo per corrispondenza – non rivolgersi a questo indirizzo per riparazioni o sostituzioni. Si pregano i clienti europei di rivolgersi al rivenditore.) 12 GARANZIA Si garantisce che il termometro a due ingressi TMD90 sarà esente da difetti di materiale e di fabbricazione per un anno a decorrere dalla data di acquisto da parte dell’acquirente o dell’utente originale. Se durante il periodo di garanzia si ritiene che il termometro sia difettoso, restituirlo, allegando lo scontrino, a un centro di assistenza Meterman Test Tools oppure al rivenditore o distributore locale Meterman Test Tools presso cui è stato acquistato. Per ulteriori informazioni vedere la sezione Riparazioni. La durata di qualsiasi garanzia implicita attivata in base alla vendita di un termometro Meterman Test Tools, incluse ma non a titolo esclusivo le garanzie implicite di commerciabilità e idoneità per uno scopo specifico, è limitata al periodo sopra specificato di un anno. La Meterman Test Tools non sarà responsabile del mancato utilizzo del termometro o di altri danni incidentali o indiretti, di spese o perdite economiche, o delle relative richieste di risarcimento. Alcuni Paesi non consentono limitazioni sulla durata delle garanzie implicite e/o l’esclusione o la limitazione di danni incidentali o indiretti; di conseguenza le precedenti limitazioni o esclusioni non saranno applicate. Questa garanzia offre specifici diritti legali ed è possibile che, secondo il Paese in cui si vive, si abbiano altri diritti. Descrizione delle termocoppie Tipo J Lega Ferro Fe K Nichel-cromo (Chromel) Ni-Cr Platino-rodio al 13% Pt-13% Rh Platino-rodio al 10% Pt-10% Rh Rame Cu R S T E Nichel-cromo (Chromel) Ni-Cr 13 Lega Rame-nichel (Constantan) Cu-Ni Nichel-alluminio Ni-AL Platino Pt Platino Pt Rame-nichel (Constantan) Cu-Ni Rame-nichel (Constantan) Cu-Ni Dati tecnici Tipo T/C K J T R/S E Portata da -200 a 650 °C da 651 a 1370 °C da -328 a 1000 °F da 1001 a 2498 °F da -200 a 500 °C da 501 a 760 °C da -328 a 940 °F da 941 a 1400 °F da -200 a 390 °C da -328 a 730 °F da 0 a 1000 °C da 1001 a 1760 °C da 32 a 1000 °F da 1001 a 3200 °F da -200 a 380 °C da 381 a 736 °C da -328 a 720 °F da 721 a 1832 °F Risoluzione Precisione 0,1 °C 1,0 °C 0,1 °F 1,0 °F (0,1% lettura + 0,7 °C) 0,1 °C 1,0 °C 0,1 °C 1,0 °F 0,1 °C 0,1 °F 0,1 °C 1,0 °C (0,1% lettura + 0,7 °C) 0,1 °F 1,0 °F (0,1% lettura +1,4 °F) 0,1 °C 1,0 °C 0,1 °F 1,0°F (0,3% lettura + 0,7 °C) (0,1% lettura +1,4 °F) (0,1% lettura + 1,4 °F) (0,1% lettura + 0,7 °C) (0,1% lettura +1,4 °F) (0,3% lettura + 0,7 °C) (0,1% lettura +1,4 °F) Dati tecnici aggiuntivi CONDIZIONI DI FUNZIONAMENTO Da 0 a 50 °C (32-122 °F) a umidità relative fra 0 e 85% (senza condensa) Coefficiente ambientale Da 0 a 18 °C e da 28 a 50 °C (temperature ambiente). Per ciascun grado centigrado di temperatura ambiente al di sotto di 18 °C o sopra 28 °C, aggiungere nelle specifiche di precisione la seguente tolleranza: 0,01% della lettura + 0,03 °C (0,01% della lettura + 0,06 °F). 14 Termómetro TMD90 de doble entrada 1 Español Información sobre seguridad................................................................2 Símbolos utilizados en este manual ...................................................2 Introducción ........................................................................................3 Pantalla y controles..............................................................................4 Funciones ............................................................................................5 Apagado automático (modo de reposo) .............................................5 Cambio del tipo de termopar .............................................................5 Uso de los botones pulsadores..........................................................6 Salida RS-232...................................................................................8 Localización y resolución de fallos........................................................9 Reemplazo de la batería .....................................................................11 Adaptador de 9 V CA ..........................................................................11 Reparación ........................................................................................12 GARANTÍA .........................................................................................13 Definiciones de los termopares...........................................................13 Especificaciones..............................................................................14 Especificaciones adicionales...............................................................14 Español Índice Información sobre seguridad • • • • • • Conecte ÚNICAMENTE termopares en la entrada de termopares. Asegúrese de que su medidor esté configurado para el tipo de termopar que se utilizará. Cerciórese de que el termopar que utilice pueda soportar la temperatura extrema a la cual estará expuesto. Mantenga adecuadamente el medidor y calíbrelo con regularidad Utilice el termómetro solamente de la manera que se especifica en este manual; de lo contrario, la protección provista por el termómetro podría verse afectada. Reemplace la batería tan pronto como aparezca BAT para evitar lecturas falsas que podrían traer como consecuencia choques eléctricos y lesiones. Advertencia El termopar suministrado no está diseñado para tener contacto con líquidos ni circuitos eléctricos con presencia de tensión. XW Símbolos utilizados en este manual X T -P Tensión peligrosa Aislamiento doble Cumple con las directivas de la Unión Europea. W J 2 Consulte el manual Conexión a tierra Introducción El modelo TMD90 es un termómetro de doble entrada que acepta termopares tipo K, J, T, R, S y E. Cuenta con una pantalla de visualización triple para ver los resultados de la medición, además de un reloj de tiempo relativo. El software y cable RS-232 opcionales permiten la recolección de datos en tiempo real en un PC, para su posterior análisis o redacción de informes. El modelo TMD90 muestra todos los segmentos LCD (pantalla de cristal líquido) durante aproximadamente 3 segundos cuando se lo enciende inicialmente, y luego proporciona información sobre cualquier termopar que se hubiera conectado. Se dispone de numerosas combinaciones de visualización. La pantalla LCD está dividida en tres secciones distintas: una pantalla superior grande (principal), y dos pantallas inferiores más pequeñas (secundaria y reloj relativo). Las tres zonas de visualización se actualizan continuamente con las mediciones de temperatura y la información sobre tiempo relativo. Existen diversas opciones referentes a cómo y cuál información se presenta en la pantalla LCD. • Las lecturas de temperatura se alternan fácilmente entre grados Fahrenheit y centígrados. • Una luz de fondo ilumina la pantalla LCD para permitir la visualización en áreas de luz tenue. congelará los datos de la pantalla superior mientras • El botón permite al mismo tiempo que las pantallas inferiores pueden continuar actualizando la información. • También se exhibe un indicador de batería con poca carga cuando corresponda. • El medidor exhibe de manera predeterminada el último modo seleccionado cuando se lo había apagado. Nota: Si no hay termopares conectados, aparecerán cuatro guiones (---) en la pantalla de temperatura. Equipo incluido Termómetro TMD90 1 Sonda de termopar tipo K 1 Manual del usuario 1 Estuche de transporte de plástico duro 1 Batería de 9 V (instalada) 1 A 3 Pantalla y controles 10 (T2) 10 (T1) 2 3 4 1 5 7 6 8 11 9 6 12 13 A B C D E F G H I J K L M REC El reloj relativo está activo para registrar las ocurrencias de temperatura mínima y máxima o para mostrar fecha y hora. MAX/MIN Se están supervisando las mediciones de temperatura mínima o máxima. HOLD/REL Congela la pantalla primaria o establece un cero relativo para la información de la pantalla principal. BAT Indicador de carga escasa de la batería. °C °F Muestra la temperatura en grados centígrados o Fahrenheit. T1-T2 Alterna la información de las pantallas desde T1 (principal)/ T 2 (secundaria) a T2 (principal) / T1 (secundaria), luego de T 1-T 2 (diferencial) en la pantalla principal y alternando las temperaturas T1/T2 en la pantalla secundaria. Pantalla principal de datos. Muestra T1, T2 o T1-T2 (diferencial de temperatura -TD ) o un cero relativo de T1, T2 o TD. KJTRSE Selecciona la referencia de entrada correcta para el termopar en uso. T1 y T2 deben ser del mismo tipo de termopar. Pantalla secundaria de datos. Muestra T2, T1 o las temperaturas T1 y T2. Entrada del termopar. Enchufes polarizados positivo y negativo para las sondas del termopar utilizadas (tipo de patilla); T1 a la izquierda, T2 a la derecha. Pantalla de reloj relativo. Exhibe el tiempo en horas, minutos y segundos (HHMMSS) al pulsarse REC, y el tiempo relativo en que fueron registrados los datos MIN o MAX. Puerto de salida RS-232. Software y cable opcionales – PN TMSW. Entrada del convertidor de CA a CC. 4 W Precaución Lea toda la información de seguridad antes de utilizar este termómetro. Funciones Apagado automático (modo de reposo) DA El medidor TMD90 se apaga automáticamente aproximadamente 20 minutos después de ser encendido. Para registrar u operar a lo largo de períodos de tiempo más prolongados, puede desactivar el modo de reposo pulsando y simultáneamente, cuando se enciende el instrumento. Cuando aparece "n" en el centro de la pantalla, puede soltar el botón On. Se desactiva la función de apagado automático al apagarse el medidor. ED E C Cambio del tipo de termopar Pulse y simultáneamente durante 2 o más segundos . El tipo de sonda pasará hasta que aparezca K. Pulse el botón cíclicamente a través de los tipos K (el tipo predeterminado), J, T, R, S y E. El modo vigente se muestra del lado izquierdo de la pantalla LCD. Pulse ; aparecerá una "S" en el centro de la pantalla. 5 Uso de los botones pulsadores La pantalla arranca de manera predeterminada en el último modo utilizado. Para su conveniencia, el medidor utiliza de manera predeterminada los ajustes empleados durante la última operación. REC HOLD REL ˚C ˚F T 1 T2 6 Para controlar la operación del TMD 90 utilice los pulsadores. La tabla supone que se ha encendido el TMD90 con dos termopares instalados y que se lo ha configurado de modo de mostrar T1 (opción predeterminada) en la pantalla principal, un termopar de tipo K y la función de registro apagada. DA Botón F B C E Descripción Enciende y apaga el medidor. Pulse el botón HOLD/REL momentáneamente y la pantalla principal (T1, T2 o T1-T2) se congela, con la palabra HOLD exhibida en la parte superior. Pulse este botón durante dos o más segundos y aparecerá REL en la parte superior de la pantalla LCD; la pantalla principal REL indica el cero relativo. El cero relativo hace que el valor de la pantalla principal se muestre como "000.0" – sólo la medida del cambio de temperatura será indicada. Se pueden registrar temperaturas relativas. Pulse HOLD/REL una vez más para que la unidad regresa a su valor predeterminado. Pulse momentáneamente y la pantalla principal cambiará a T2 (la pantalla secundaria muestra T1); pulse este botón mo-mentáneamente una vez más y mostrará T1-T2; la pantalla secundaria alterna entre T1 y T2; pulse momentáneamente una vez más y el instrumento retornará a su valor predeterminado. Pulse este botón momentáneamente y la luz de fondo se iluminará durante aproximadamente 30 segundos, para luego apagarse. Pulse este botón momentáneamente y la unidad alternará entre las temperaturas en Fahrenheit (la opción predeterminada) y en grados centígrados. El modo vigente está indicado del lado derecho de la pantalla LCD. Pulse este botón momentáneamente y el reloj relativo comenzará a funcionar en la pantalla inferior derecha; aparecerá del lado izquierdo superior de la pantalla LCD – Todas las demás funciones del botón quedarán bloqueadas, salvo Power y Backlight. T1, T2 o T1-T2 se muestran en la pantalla primariaprincipal; la pantalla secundaria continúa actualizándose. Pulse momentáneamente una vez más y la unidad pasará cíclicamente a través de las temperaturas MAX y MIN (temperaturas máxima y mínima registardas) y regresará a la temperatura actual. El modo de registro se muestra en la pantalla LCD. Pulse y mantenga pulsado este botón durante tres segundos para desactivar la función de registro. 7 Salida RS-232 Con el software y cable opcionales, P/N TM-SW, el TMD90 puede enviar los resultados de las mediciones a una computadora que cuente con alguno de los siguientes sistemas operativos: Windows 95/98/NT/2000/XP/ME. Características del software: • Registra hasta 16,000 lecturas de muestra con memoria del reloj en tiempo real. • Intervalo de muestras programable desde 1 seg hasta 3600 seg. • Configuraciones de alarma máxima y mínima definidos por el usuario. • Permite un posterior análisis mediante la descarga de los datos a un programa informático. • Todos los datos se guardan en formato .txt para su fácil transformación en otro programa de software. 8 Localización y resolución de fallos Problema El termómetro no se enciende. Solución • Verifique el voltaje de la batería y reemplace la misma si está descargada. • Verifique que los clips de la batería sujeten de manera firme los bornes de la misma. • Inserte el termopar faltante. Aparecen guiones en las pantallas de datos T1 y T2. • Mida la resistencia de los termopares para asegurarse de que no estén partidos internamente. • Limpie el termopar y reinicie el proceso. Aparecen guiones en las pantallas de datos T1 y/o T2, con los termopares insertados. 9 Problema Tienen lugar desplazamientos de temperatura desde un valor conocido en un entorno controlado E El reloj relativo no se inicia cuando se pulsa el botón . Aparecen guiones durante una revisión del valor máximo registrado. Solución • Verifique que el tipo de termopar coincida con el icono mostrado. • Limpie y seque las patillas del termopar y permita que se sequen al aire. • Confirme la temperatura con un termopar que se sepa que está en buenas condiciones. En caso necesario reemplace el termopar defectuoso. • Verifique que el termopar se haya insertado correctamente. • El termopar no se reconoce. Compruebe que no haya una desconexión intermitente o momentánea del termopar. A DA Los datos continúan actualizándose después de que usted pulsa el botón . A • Verifique que aparezca el icono HOLD en la pantalla. Pulse el botón con firmeza. El instrumento se apaga mientras se encuentra registrando. • Pulse y simultáneamente antes del encendido para desactivar la característica de apagado automático. 10 Reemplazo de la batería Reemplace la batería de 9 V cuando: • Aparece BAT del lado derecho de la pantalla. • El medidor no se enciende. • Aparece BAT cuando usted enciende la luz de fondo. Incluso si se reemplazó la batería recientemente, verifique el nivel de voltaje en caso de no haber respuesta del medidor. Para reemplazar la batería 1. Retire todos los termopares de la parte superior del medidor. 2. Coloque el instrumento boca abajo sobre una superficie plana y limpia. 3. Afloje el tornillo y retire la puerta del compartimiento de la batería. 4. Reemplace la batería. 5. Instale nuevamente la puerta del compartimiento de la batería y apriete el tornillo. Retire la batería si no tiene intención de utilizar el medidor durante un mes o más. No deje la batería en un medidor que pueda ser expuesto a extremos de temperatura. Adaptador de 9 V CA Para el registro de mediciones a largo plazo, se puede usar un adaptador de 9 V CA en lugar de la batería. Dicho adaptador se puede adquirir en cualquier tienda de artículos electrónicos, ajustándose a las siguientes especificaciones: Enchufe eléctrico en miniatura con positivo y conductor externo negativo, con un valor nominal de corriente de 500 mA. Cuando se utiliza el adaptador de CA, la batería de 9V puede permanecer en el TMD90. 11 Reparación Todas las herramientas de prueba devueltas para su calibración o reparación, cubiertas o no por la garantía, deberán estar acompañadas por lo siguiente: su nombre, el nombre de la empresa, la dirección, el número de teléfono y una prueba de compra. Además, incluya una breve descripción del problema o del servicio solicitado y las puntas de prueba del termómetro. Los pagos correspondientes a reparaciones o reemplazos no cubiertos por la garantía se deben remitir a la orden de Meterman Test Tools en forma de cheque, giro postal, pago mediante tarjeta de crédito (incluir el número y la fecha de vencimiento) u orden de compra. Reparaciones y reemplazos cubiertos por la garantía (todos los países) Antes de solicitar una reparación sírvase leer la siguiente declaración de garantía y compruebe el estado de las baterías y los fusibles de su medidor. Durante el periodo de garantía, toda herramienta de comprobación en mal estado de funcionamiento puede ser devuelta al distribuidor de Meterman Test Tools para cambiarla por otra igual o un producto semejante. Consulte la sección “Dónde comprar” del sitio www.metermantesttools.com en Internet para obtener una lista de los distribuidores de su zona. Además, en los Estados Unidos y Canadá, las unidades para reparación y reemplazo cubiertas por la garantía también se pueden enviar a un Centro de Servicio de Meterman Test Tools (las direcciones se incluyen más adelante). Reparaciones y reemplazos no cubiertos por la garantía (Estados Unidos y Canadá) Las unidades para reparaciones no cubiertas por la garantía en Estados Unidos y Canadá se deben enviar a un Centro de Servicio de Meterman Test Tools. Póngase en contacto con Meterman Test Tools o con el vendedor de su producto para solicitar información acerca de los precios vigentes para reparación y reemplazo. En Estados Unidos En Canadá Meterman Test Tools Meterman Test Tools 1420 75th Street SW 400 Britannia Rd. E. Unit #1 Everett, WA 98203 Mississauga, Ontario L4Z 1X9 Tel.: 888-993-5853 Tel.: 905-890-7600 Fax: 425-446-6390 Fax: 905-890-6866 Reparaciones y reemplazos no cubiertos por la garantía – Europa El distribuidor de Meterman Test Tools puede reemplazar aplicando un cargo nominal las unidades vendidas en Europa no cubiertas por la garantía. Consulte la sección “Dónde comprar” del sitio www.metermantesttools.com en Internet para obtener una lista de los distribuidores de su zona. Dirección para envío de correspondencia en Europa* Meterman Test Tools Europe P.O. Box 1186 5602 B.D. Eindhoven Holanda *(Correspondencia solamente. En esta dirección no se suministran reparaciones ni reemplazos. Los clientes europeos deben ponerse en contacto con el distribuidor). 12 GARANTÍA El termómetro TMD90 Dual Input está garantizado contra defectos en los materiales o la mano de obra por un periodo de un (1) año a partir de la fecha de compra del mismo por el comprador o usuario original. Todo termómetro considerado defectuoso durante el periodo de garantía debe ser devuelto con la correspondiente prueba de compra a un Centro de Servicio autorizado de Meterman Test Tools o al representante o distribuidor local en donde fue adquirido el instrumento. Consulte la sección Reparación para obtener información más detallada. La duración de todas las garantías implícitas resultantes de la venta de un termómetro Meterman Test Tools, incluidas pero no limitadas a las garantías de comerciabilidad e idoneidad para un propósito determinado, se limitan al periodo de garantía de un (1) año establecido anteriormente. Meterman Test Tools no se responsabiliza por la imposibilidad de utilizar el termómetro u otros perjuicios, gastos o pérdidas económicas consecuentes o incidentales, ni por ningún tipo de reclamo o reclamos por tales perjuicios, gastos o pérdidas económicas. Ya que algunos estados no permiten la limitación de la duración de una garantía implícita, ni la exclusión ni limitación de daños incidentales o consecuentes, las limitaciones y exclusiones anteriores pueden no regir para usted. Esta garantía le otorga derechos legales específicos y es posible que también tenga otros derechos que varían de un estado a otro. Definiciones de los termopares Tipo J K R S T E Aleación Hierro Fe Níquel-Cromo (Chromel) Ni-Cr Platino-13% Rodio Pt-13% Rh Platino-10% Rodio Pt-10% Rh Cobre Cu Níquel-Cromo (Chromel) Ni-Cr 13 Aleación Cobre-Níquel (Constantan) Cu-Ni Níquel-Aluminio Ni-AL Platino Pt Platino Pt Cobre-Níquel (Constantan) Cu-Ni Cobre-Níquel (Constantan) Cu-Ni Especificaciones Tipo de termopar K J T R/S E Rango -200 a 650 °C 651 a 1370 °C -328 a 1000 °F 1001 a 2498 °F -200 a 500 °C 501 a 760 °C -328 a 940 °F 941 a 1400 °F -200 a 390 °C -328 a 730 °F 0 a 1000 °C 1001 a 1760 °C 32 a 1000 °F 1001 a 3200 °F -200 a 380 °C 381 a 736 °C -328 a 720 °F 721 a 1832 °F Resolución 0,1 °C 1,0 °C 0,1 °F 1,0 °F 0,1 °C 1,0 °C 0,1 °C 1,0 °F 0,1 °C 0,1 °F 0,1 °C 1,0 °C 0,1 °F 1,0 °F 0,1 °C 1,0 °C 0,1 °F 1,0°F Precisión (0,1% lectura + 0,7 °C) (0,1% lectura + 1,4 °F) (0,1% lectura + 0,7 °C) (0,1% lectura + 1,4 °F (0,1% lectura + 0,7 °C) (0,1% lectura + 1,4 °F) (0,3% lectura + 0,7 °C) (0,1% lectura + 1,4 °F) (0,3% lectura + 0,7 °C) (0,1% lectura + 1,4 °F) Especificaciones adicionales CONDICIONES DE OPERACIÓN 0 a 50 °C (32 a 122 °F) a valores de humedad relativa del 0 al 85 % (sin condensación) Coeficiente ambiente 0 a 18 °C y 28 a 50 °C (temperaturas ambiente) Para cada °C de temperatura ambiente menor que 18 °C o mayor que 28 °C, agregue la siguiente tolerancia a la especificación de exactitud: 0,01% de la lectura + 0,03 °C (0,01% de la lectura + 0,06 °F) 14 TMD90 TMD90 ® ® ® ® U.S. Service Center Canadian Service Center Dual Input Digital Thermometer European Correspondence Address* User Manual Meterman Test Tools 1420 75th Street SW Everett, WA 98203 Tel: 888-993-5853 Fax: 425-446-6390 Meterman Test Tools 400 Britannia Rd. E. Unit #1 Mississauga, ON L4Z 1X9 Tel: 905-890-7600 Fax: 905-89-6866 Meterman Test Tools Europe P.O. Box 1186 5602 BD Eindhoven The Netherlands *Correspondence only - no repair or replacement available from this address. European customers please contact your distributor. Visit www.metermantesttools.com for • Catalog • Application notes • Product specifications • Product manuals PN 2099366 September 2003 © Wavetek Meterman Test Tools. All rights reserved. Printed in China. Please Recycle • Mode d'emploi • Bedienungshandbuch • Manuale d'Uso • Manual de uso
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70

Wavetek 90 Manual de usuario

Tipo
Manual de usuario
Este manual también es adecuado para