Ohaus ST3100-B Manual de usuario

Categoría
Medir, probar
Tipo
Manual de usuario

Este manual también es adecuado para

STARTER 3100
Bench pH Meter
Instruction Manual
Manual de instrucciones
del pHmetro de mesa
STARTER 3100
STARTER 3100
PH-mètre de laboratoire
Manuel d’instructions
STARTER 3100
Medidor de pH de Bancada
Manual de Instruções
TABLE OF CONTENTS
1 INTRODUCTION ................................................................... 1
1.1 Definition of Signal Warnings and Symbols ...................................... 1
1.2 Safety Precautions ............................................................................ 2
1.3 Display and controls .......................................................................... 3
2 INSTALLATION ..................................................................... 5
2.1 Package contents .............................................................................. 5
2.2 Installing the stand-alone electrode holder ....................................... 6
2.3 Installing the power adapter .............................................................. 6
2.4 Connect the pH electrodes ................................................................ 7
2.5 Meter stand for adjusting view angle ................................................. 7
2.6 Attached quick guide ......................................................................... 7
3 SETUP ................................................................................... 8
3.1 Set temperature unit and MTC value ................................................ 8
3.2 Selecting a predefined buffer group .................................................. 8
4 STARTER 3100 OPERATION ............................................... 9
4.1 Calibration ......................................................................................... 9
4.1.1 Buffer group ................................................................................... 9
4.1.2 Performing 1-point calibration ..................................................... 10
4.1.3 Performing 2-point calibration ...................................................... 11
4.2 Sample measurement ..................................................................... 12
4.2.1 pH measurement ......................................................................... 12
4.2.2 mV measurement ........................................................................ 12
4.3 Temperature measurement ............................................................. 12
4.4 Using the memory ........................................................................... 12
4.4.1 Storing a reading ......................................................................... 12
4.4.2 Recalling from memory ............................................................... 12
4.4.3 Clearing the memory ................................................................... 13
4.5 Printing ............................................................................................ 13
5 MAINTENANCE .................................................................. 15
5.1 Error message ................................................................................. 15
5.2 Meter maintenance.......................................................................... 15
5.3 Electrode maintenance ................................................................... 16
5.4 Self-diagnosis .................................................................................. 16
5.5 Recover factory settings .................................................................. 16
6 TECHNICAL DATA .............................................................. 17
6.1 Specifications .................................................................................. 17
6.2 Compliance ..................................................................................... 18
7 BUFFER GROUP ................................................................ 19
STARTER 3100 Bench pH Meter EN-1
1 INTRODUCTION
Thank you for choosing OHAUS.
Please read the manual completely before using the STARTER 3100 bench pH meter to ensure
proper setup, operation and maintenance.
STARTER 3100 has an excellent performance/price ratio and is designed with many useful features
including a stand-alone electrode holder and a beeper for error alert. Other accessories such as
pH electrodes and buffer solutions are also available.
STARTER 3100 offers many practical features and functions such as:
Large backlit LCD with Quick Guide attached under the meter to assist with user operation
Stand-alone electrode holder for maximum flexibility
Electrode condition icon automatically showing you the health of your pH electrode
RS232 port and memory to store up to 99 measurements
1.1 Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and
warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument,
malfunctions and false results.
Signal Words
WARNING For a hazardous situation with medium risk, possibly resulting in
injuries or death if not avoided.
CAUTION For a hazardous situation with low risk, resulting in damage to the
device or the property or in loss of data, or injuries if not avoided.
Attention For important information about the product.
Note For useful information about the product
Warning Symbols
General hazard
Explosion hazard
Corrosive hazard
Alternating current
Direct current
EN-2 STARTER 3100 Bench pH Meter
1.2 Safety Precautions
CAUTION: Read all safety warnings before installing, making connections, or servicing this
equipment. Failure to comply with these warnings could result in personal injury and/or property
damage. Retain all instructions for future reference.
Verify that the input voltage range printed on the data label and the plug type matches the
local AC power to be used.
Make sure that the power cord does not pose a potential obstacle or tripping hazard.
Use the equipment only in dry locations.
Dry off any liquid spills immediately. The instrument is not watertight.
When using chemicals and solvents, comply with the instructions of the chemical producer
and the general lab safety rules.
Use only approved accessories and peripherals.
Operate the equipment only under ambient conditions specified in these instructions.
Disconnect the equipment from the power supply when cleaning.
Do not operate the equipment in hazardous or unstable environments.
Service should only be performed by authorized personnel.
STARTER 3100 Bench pH Meter EN-3
1.3 Display and controls
Displays
1
Electrode condition
Slope: more than 95%
and offset: ± (0-15) mV
Electrode condition is good
Slope: less than 90%
or offset: ± (35-60) mV
Electrode condition is not
good or needs cleaning
2
Endpoint stability icon ; Auto endpoint icon
3
Measurement icon - ; means measurement or calibration is running when blinks
4
Calibration icon - ; means calibration in progress when display
5
Setup icon - ; instrument is in the setup mode, can set temperature(MTC), buffer group
etc.
6
pH/mV reading or slope in calibration process
7
Calibration point / Buffer group /Memory number / Error index
8
Auto temperature compensation - ATC ; Manual temperature compensation - MTC
9
Temperature during measurement or offset (mV) in calibration process
2
3
4
1
5
7
8
9
1
2
3
4
5
6
EN-4 STARTER 3100 Bench pH Meter
Controls
Button
Press & release
Press & hold
for 3 seconds
- Start or finish measurement
- Confirm setting, store entered
value
- Turn auto endpoint on / off
,
- Start calibration
- Recall the latest calibration
data : slope and offset
- Meter turn on
- Exit and return to measurement
screen
- Meter turn off
- Store current reading to memory
- Increase value during setting
- Scroll up through the memory
- Recall stored data
- Print current memory data
- Switch between pH and mV
- Decrease value during setting
- Scroll down through the memory
- Enter setup mode
- Start self-diagnosis
Turn on/turn off the backlight
of the LCD
STARTER 3100 Bench pH Meter EN-5
2 INSTALLATION
Carefully unpack the box.
2.1 Package contents
The model ST3100-B (basic package) has the following items:
In addition to ST3100-B content, the ST3100-F package also includes the following:
In addition to ST3100-B content, the ST3100-H package also includes the following:
Each pH buffer powder should be dissolved in 250ml pure water or deionized water in a
volumetric flask.
Additional electrodes available include the following:
Model
Description
P/N
ST310
3-in-1 plastic refillable pH Electrode
83033965
ST210
2-in-1 plastic refillable pH Electrode
83033966
ST320
3-in-1 plastic gel pH Electrode(no need to fill)
83033967
ST230
2-in 1 glass muddy sample pH Electrode
83033968
STORP1
Gel plastic ORP electrode
30038555
STORP2
Refillable glass ORP electrode
30038553
STTEMP30
Temperature Electrode
83033970
ST3100-B
Units
STARTER 3100
1
Stand-alone electrode holder
1 set
In use cover
1
12V Power supply
1 set
pH Buffer Powder Set (4.01, 7.00, 10.01)
1 set
ST310 3-in-1 refillable pH electrode
1
pH Buffer Powder Set (4.01, 7.00, 10.01)
1 set
ST320 3-in-1 gel pH electrode
1
EN-6 STARTER 3100 Bench pH Meter
Buffers and Accessories:
Buffer powder set (4.01; 7.00; 10.00)
83033971
Buffer pH1.68 250ml
30100424
Buffer pH4.01 250ml
30100425
Buffer pH7.00 250ml
30100427
Buffer pH10.01 250ml
30100429
Buffer pH12.45 250ml
30100430
Electrode holder stand alone (3100/3100C)
30058733
In use cover
30058734
2.2 Installing the stand-alone electrode holder
Install the electrode arm on the base,
After adjusting the tension knob to some extent, you can move the upper arm up and down.
When the pH electrode is installed on the arm, the storage bottle of pH electrode fits into the
base hole.
2.3 Installing the power adapter
Insert the right adapter clip into the power adapter slot. The Ohaus balance power supply is also
12V which could also be used for the pH meter.
STARTER 3100 Bench pH Meter EN-7
2.4 Connect the pH electrodes
There are 3 sockets for electrode. pH socket (BNC); TEMP. Socket (Cinch) and REF Socket
(2mm banana).
For ST310 electrode connect to the BNC and Cinch socket. For 2-in-1 pH electrode (e.g. ST210)
you only need to connect the BNC.
REF socket is for separate reference electrode. (e.g.STREF1)
2.5 Meter stand for adjusting view angle
One unique design of the STARTER 3100 is the meter stand, which can be used to adjust the view
angle of the display in case of working on high table.
2.6 Attached quick guide
Another unique design of the STARTER 3100 is the attached quick guide, the quick guide is
attached into the bottom housing of the meter.
EN-8 STARTER 3100 Bench pH Meter
3 SETUP
3.1 Set temperature unit and MTC value
Please note:
If a temperature electrode is used, Automatic Temperature Compensation (ATC) and the sample
temperature are displayed on the screen. You may then choose to skip MTC setup (below).
If the meter does not detect a temperature electrode or one is not used, the meter automatically
switches to Manual Temperature Compensation (MTC) mode and MTC appears on the screen.
MTC can be set as follows:
Power the meter on by pressing .
Press and hold until the setup icon appears on the display and the current
temperature unit blinks (°C or °F).
Press or to switch between °C and °F.
Press to confirm your selection.
Then
Continue with MTC temperature setting by using or to adjust temperature
compensation accordingly
Press to confirm the setting
Press to return to the measurement screen.
The default MTC temperature value setting is 25°C (77°F).
Note: °C = 5/9 (°F - 32)
3.2 Selecting a predefined buffer group
After confirming the MTC compensation temperature value, make the buffer group selection. Use
or to select a buffer group among 3 buffer groups (4.1.1). Press to
confirm the setting or press to leave to return to the measurement screen.
The default buffer group is b1.
b1
1.68
4.01
7.00
10.01
STARTER 3100 Bench pH Meter EN-9
4 STARTER 3100 OPERATION
Standard procedure of pH measurement is as follows:
a) pH electrode preparation
b) buffer preparation and pH electrode calibration
c) sample preparation
d) pH measurement
e) Record measurement results or print
f) Rinse the pH electrode and properly store
pH electrode preparation: pH electrode should be rinsed with pure water before and after using.
Check if the electrode is physically damaged. (Be careful with the glass bulb)
The pH electrode should be stored in the storage bottle; the solution in the bottle is 3M KCl
solution. After placing the pH electrode into the sample or buffer solution, user should stir several
seconds then wait 30 to 60 seconds for the signal to stabilize, and then press the button to
operate (Calibration or measurement).
WARNING Do not operate the equipment in hazardous environments. The
equipment is not explosion protected.
WARNING When using chemicals and solvents, comply with the instructions of the
chemical producer and the general lab safety rules.
4.1 Calibration
4.1.1 Buffer group
STARTER 3100 can perform 1-, 2- or 3- point calibrations.
There are 3 buffer groups in the meter, you can select the buffer group you prefer (see 4.3),
default buffer is b1 US standards; the buffer value will be automatically recognized during
calibration. The 3 predefined buffer groups are (at 25°C):
STARTER 3100 automatically corrects for the temperature dependence of the buffer pH values
given in the following table - buffer group b1.
This means, if the buffer solutions’ temperature is 15°C, the calibration value you will get should
be pH1.67 (buffer 1.68), pH4.00 (buffer 4.01), pH7.04(buffer 7.00) and pH10.12 (buffer 10.01).
b1
1.68
4.01
7.00
10.01
b2
2.00
4.01
7.00
9.21
11.00
b3
1.68
4.00
6.86
9.18
12.46
EN-10 STARTER 3100 Bench pH Meter
Calibration: pH electrodes need to be calibrated with pH standard buffer solution before a proper
pH measurement can be made. Calibration is to display the right pH value when meter receive
the mV value signal from the pH electrode.
Slope: the linear coefficient between mV and pH according to theoretical value (e.g. -
59.16mV/pH @ 25 °C which means 100% slope);
Offset: the mV value when pH value is 7.00. (Theoretical value is 0 mV);
4.1.2 Performing 1-point calibration
When performing calibration, Ohaus recommends using Auto End Point Mode. After powering
the meter on, be sure the top of the screen shows to ensure the meter is in Auto End point
Mode.
Auto or Manual End point Mode:
o Press and hold to change the End Point Mode.
o When in Manual Mode, to manually reach a pH measurement or calibration value,
you need to press button- when reading is stable and displays :
then the sample reading or calibration value freezes, blinks 3 times and
freezes on the display.
o When in Auto End Point Mode, the meter determines when the reading is stable
then displays and locks the reading or calibration value automatically, the reading
freezes and blinks 3 times then disappears; blinks 3 times and freezes
on the display.
Note: With the 1-point calibration only the offset is adjusted. If the sensor was previously
calibrated with multi-point calibration the previously stored slope will remain. Otherwise theoretical
100% slope (-59.16 mV / pH) will be used.
5 °C
1.67
4.01
7.09
10.25
10 °C
1.67
4.00
7.06
10.18
15 °C
1.67
4.00
7.04
10.12
20 °C
1.68
4.00
7.02
10.06
25 °C
1.68
4.01
7.00
10.01
30 °C
1.68
4.01
6.99
9.97
35 °C
1.69
4.02
6.98
9.93
40 °C
1.69
4.03
6.97
9.89
45 °C
1.70
4.05
6.97
9.86
50 °C
1.71
4.06
6.96
9.83
STARTER 3100 Bench pH Meter EN-11
When STARTER 3100 is in pH measurement mode (see 4.2); place the pH electrode in a
calibration buffer, stir 5 seconds, wait for 30 seconds, then:
Press Cal 1displays on the bottom left of the screen and Cal is blinking.
and appear on the top of the screen, is blinking during calibration.
The meter reaches endpoint automatically according to the preselected auto-endpoint
mode, the calibration point pH value (e.g. 7.00) with the temperature display on the screen.
The 1-point calibration is finished; There are now 3 options (OHAUS recommends conducting at
least a 2 point calibration);
1. Press to do the 2-point calibration.
2. Press to store the 1-point calibration and exit, the offset and the slope are
shown on the display for 3 seconds then return to the measurement screen.
3. Press to reject the calibration, return to the measurement screen.
4.1.3 Performing 2-point calibration
Perform 1-point calibration as described above.
Rinse the pH electrode with pure water and wipe off with tissue.
Place the electrode in the next calibration buffer, stir and wait, then press , Cal 2
displays on the bottom left of the screen and Cal is blinking. On the top of the screen
and appear, is blinking during calibration.
The meter reaches endpoint according to the endpoint mode, the calibration point pH value
(e.g. 4.01) with the temperature display on the screen.
The 2-point calibration is finished. There are 3 options:
1. Press to do the 3-point calibration.
2. Press to store the 2-point calibration and exit, the offset and slope are
shown on the display for 3 seconds then return to the measurement screen.
3. Press to reject the calibration, return to the measurement screen.
Note: To perform a 3 point calibration, follow the instructions for performing a 2 point calibration
using a third buffer.
EN-12 STARTER 3100 Bench pH Meter
4.2 Sample measurement
4.2.1 pH measurement
Place the pH electrode in the sample. Stir for 5 seconds; wait for 30 seconds.
Press to start the pH measurement, appears on the display. is blinking during
measurement.
When meter reaches endpoint, the pH value with the temperature display on the screen.
4.2.2 mV measurement
Press to switch between pH measurement mode and mV measurement mode.
Follow the same procedure as for pH measurement to perform mV measurement.
4.3 Temperature measurement
For better accuracy, we recommend to use either a built-in or a separate temperature electrode.
If a temperature electrode is used, ATC and the sample temperature are displayed.
If the meter does not detect a temperature electrode, it automatically switches to the manual
temperature compensation mode and MTC appears. MTC temperature should be set.
Note: STARTER 3100 accepts NTC 30 temperature sensor. ATC (Auto Temperature
Compensation) or MTC (Manual Temperature Compensation), the Temperature Compensation
only corrects for the change in the output of the electrode, not for the change in the actual solution.
This means the meter corrects the signal(mV) from the electrode to get a more accurate pH value
according to the real temperature.
4.4 Using the memory
4.4.1 Storing a reading
The STARTER 3100 can store up to 99 endpoint results.
Press button when the measurement reaches endpoint. M01 indicates that one result
has been stored.
If you press when M99 is displayed, FUL displays to indicate the memory is full. To store
further data you will have to clear the memory. (See 4.4.3)
4.4.2 Recalling from memory
Press and hold to recall the stored values from memory when the current
measurement reaches endpoint.
STARTER 3100 Bench pH Meter EN-13
Press button or to scroll through the stored results. R01 to R99 indicates which
result is being displayed.
Press to exit.
4.4.3 Clearing the memory
Pressing or to scroll through the stored results until MRCL appears.
Press , CLr blinks;
There are now 2 options:
Press to confirm the deletion of all the stored data.
Press to return to the measurement mode without deleting the memory.
4.5 Printing
If the printer is connected to the STARTER 3100(e.g. SF-F40A, green light means connection
success; printer baud rate 9600bps; 8 data bit; none parity bit; 1 stop bit), a print-out is
automatically generated after each end pointed measurement or calibration.
The format for the print-out following a pH measurement is:
End Point, Value, Temp., ATC/MTC
Auto EP, 4.01pH, 25.0 °C, MTC
The details for the second line are:
Auto EP, 4.01pH, 25.0 °C , MTC
| | | |---- Manual Temperature Compensation
| | |---------------- Temperature value and unit ºC
| |------------------------- pH value
|---------------------------------------- Auto End Point
The format for the print-out following a mV measurement is:
End Point, Value, Temp., ATC/MTC
Manual EP, 182mV, 23.2 °C, ATC
The print-out in case of an error message is:
End Point, Value, Temp., ATC/MTC
Error3
The print-out for a 2-point calibration is:
Buffer1: 4.01pH
mV1: 178mV
Temp.1: 25.0 °C
Buffer2: 7.00pH
mV2: 3mV
Temp.2: 25.0 °C
EN-14 STARTER 3100 Bench pH Meter
Slope: 99%
Offset: 5mV
ATC/MTC: MTC
The print-out for a 3-point calibration is:
Buffer1: 4.01pH
mV1: 178mV
Temp.1: 25.0 °C
Buffer2: 7.00pH
mV2: 0mV
Temp.2: 25.0 °C
Buffer3: 9.21pH
mV3: -130mV
Temp.3: 25.0 °C
Slope: 100%
Offset: 0mV
ATC/MTC: ATC
Printing from memory: When scrolling through the memory you can print the entry that is currently
viewed by pressing and holding (3 seconds) the . The printout format is as follows:
Recall 01:
End Point, Value, Temp., ATC/MTC
Manual EP, 4.01pH, 35.6 °C, ATC
STARTER 3100 Bench pH Meter EN-15
5 MAINTENANCE
5.1 Error message
Error 0
Memory access error
Reset to factory settings
Error 1
Self-diagnosis failed
Repeat the self-diagnosis procedure and make
sure that you finish pressing all five keys within
two minutes.
Error 2
Measured values out of range
Check if the electrode is properly connected and
placed in the sample solution.
Error 3
Measured buffer temperature
out of range (<5 or >40 °C)
Keep the pH buffer temperature within the range
for calibration
Error 4
Offset out of range
offset > 60mV or < - 60 mV
Make sure the pH buffer is correct and fresh;
Clean or replace the pH electrode.
Error 5
Slope out of range
Make sure the buffer is correct and fresh; Clean
or replace the pH electrode.
Error 6
Meter cannot recognize the
buffer
Make sure the buffer is correct and fresh; check if
the buffer has not been used more than once.
Error 9
The current data set has
already been stored once
An endpoint reading can only be stored once.
Perform a new measurement to store.
Error 10
The sample temperature out
of range
Check the sample temperature, the temperature
sensor.
If an error happens, the meter will also beep 3 times to alert.
For further technical support please contact Ohaus. (US please contact 1-800-672-7722).
5.2 Meter maintenance
Never unscrew the two halves of the housing!
The STARTER 3100 series instruments do not require any maintenance other than occasional wipe
with a damp cloth and the replacement of depleted batteries.
The housing is made of acrylonitrile butadiene styrene (ABS). This material is attacked by some
organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage should be
immediately wiped off.
EN-16 STARTER 3100 Bench pH Meter
5.3 Electrode maintenance
Make sure the electrode is filled with electrolyte solution. Always store the electrode according to
the electrode instruction manuals and do not allow it to dry out.
If the electrode response becomes sluggish or the slope is not acceptable, try the following:
o Soak the electrode in 0.1M HCl for at least 8 hours.
o For fat or oil contaminant, degrease the membrane with cotton wool soaked in either
acetone or a soap solution.
After electrode treatment, a new calibration should be performed. If the electrode slope is still not
acceptable, the electrode might need to be replaced.
5.4 Self-diagnosis
When the meter is on, press and hold and simultaneously until the
meter displays the full screen. Each icon blinks one after the other.
This way you may check whether all icons are correctly shown. The next step is to check that the
keys are functioning correctly. This requires user interaction.
When b blinks, five icons are displayed.
Press the five keys in any order. Each time you press a key an icon disappears from the
screen, continue to press the other keys until all the icons have disappeared.
When the self-diagnosis has been completed successfully, PAS appears. If self-diagnosis fails,
error message Err 1 appears.
Note: You have to finish pressing all five keys within 2 minutes, otherwise Err 1 appears and you
will have to repeat the procedure.
5.5 Recover factory settings
When the meter is off, press and hold & & together for 3
seconds, the screen displays and blinks, this means “Reset”. There are 2 options:
Press to reset factory settings (MTC, slope and offset, etc.), display
then restart the meter.
Or press to quit the setting, display then turn off the meter.
STARTER 3100 Bench pH Meter EN-17
6 TECHNICAL DATA
6.1 Specifications
Ambient conditions
Indoor use only
Altitude: Up to 2000 m
Specified Temperature range: 5 °C to 40 °C
Humidity: maximum relative humidity 80 % for temperatures up to 30 °C decreasing linearly to
50% relative humidity at 40 °C
Mains supply voltage fluctuations: up to ±10% of the nominal voltage
Installation category II
Pollution degree: 2
Operability is assured at ambient temperatures between 5 °C to 40 °C
Model
ST3100
Measuring range
-2.00…16.00 pH
–1999…1999 mV
-5 °C…110 °C
Resolution
0.01 pH
1 mV
0.1 °C
Error limits
± 0.01 pH
± 1 mV
± 0.5 °C
Calibration
3 points
3 predefined buffer groups
Memory
99 measurements
The last calibration data
Power supply
AC Adapter Input: 100-240V ~ X.XA 50/60 Hz
AC Adapter Output: 12V X.XA
Size/weight
220 W x 175 D x 78 H mm / 0.75 kg
Display
Liquid crystal with backlight
Input
BNC, impedance > 10e+12 Ω
Cinch, NTC 30 kΩ
Reference input
2mm banana socket
Temperature-compensation
ATC & MTC
Housing
ABS
EN-18 STARTER 3100 Bench pH Meter
6.2 Compliance
This product conforms to the EMC Directive 2004/108/EC and the Low Voltage
Directive 2006/95/EC. The Declaration of Conformity is available online at
europe.ohaus.com/europe/en/home/support/compliance.aspx.
In conformance with the European Directive 2002/96/EC on Waste Electrical
and Electronic Equipment (WEEE) this device may not be disposed of in
domestic waste. This also applies to countries outside the EU, per their specific
requirements. Please dispose of this product in accordance with local
regulations at the collecting point specified for electrical and electronic
equipment. If you have any questions, please contact the responsible authority
or the distributor from which you purchased this device. Should this device be
passed on to other parties (for private or professional use), the content of this
regulation must also be related. Disposal instructions in Europe are available
online at europe.ohaus.com/europe/en/home/support/weee.aspx. Thank you for
your contribution to environmental protection.
FCC Note: This equipment has been tested and found to comply with the limits for a Class
A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference in which case the user will be required to correct
the interference at his own expense.
ISO 9001 Registration
In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO 9001 by
Bureau Veritus Quality International (BVQI), confirming that the OHAUS quality
management system is compliant with the ISO 9001 standards requirements. On June 21,
2012, OHAUS Corporation, USA, was re-registered to the ISO 9001:2008 standard.
STARTER 3100 Bench pH Meter EN-19
7 BUFFER GROUP
STARTER 3100 automatically correct for the temperature dependence of the buffer group pH value
given in the following tables (b2, b3), you can find b1. (see 3.1.1)
Buffer group b2 Europe standard
Temp °C pH2.00 pH4.01 pH7.00 pH9.21 pH11.00
5 2.02 4.01 7.09 9.45 11.72
10 2.01 4.00 7.06 9.38 11.54
15 2.00 4.00 7.04 9.32 11.36
20 2.00 4.00 7.02 9.26 11.18
25 2.00 4.01 7.00 9.21 11.00
30 1.99 4.01 6.99 9.16 10.82
35 1.99 4.02 6.98 9.11 10.64
40 1.98 4.03 6.97 9.06 10.46
45 1.98 4.04 6.97 9.03 10.28
50 1.98 4.06 6.97 8.99 10.10
Buffer group b3 JJG119
Temp °C pH1.68 pH4.00 pH6.86 pH9.18 pH12.46
5 1.67 4.00 6.95 9.39 13.21
10 1.67 4.00 6.92 9.33 13.01
15 1.67 4.00 6.90 9.28 12.82
20 1.68 4.00 6.88 9.23 12.64
25 1.68 4.00 6.86 9.18 12.46
30 1.68 4.01 6.85 9.14 12.29
35 1.69 4.02 6.84 9.11 12.13
40 1.69 4.03 6.84 9.07 11.98
45 1.70 4.04 6.83 9.04 11.83
50 1.71 4.06 6.83 9.02 11.70
EN-20 STARTER 3100 Bench pH Meter
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of
delivery through the duration of the warranty period. During the warranty period Ohaus will repair,
or, at its option, replace any component(s) that proves to be defective at no charge, provided that
the product is returned, freight prepaid, to Ohaus.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to
radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or
as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty
registration card, the warranty period shall begin on the date of shipment to the authorized dealer.
No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not
be liable for any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or
your local Ohaus dealer for further details.
Índice
1 Introducción ......................................................................... 1
1.1 Definición de palabras de advertencia y símbolos de advertencia ........ 1
1.2 Precauciones de seguridad ....................................................................... 2
1.3 Pantalla y controles .................................................................................... 3
2 Instalación ............................................................................ 5
2.1 Contenido del paquete ............................................................................... 5
2.2 Instalación del soporte para electrodo autónomo ................................... 6
2.3 Instalación del adaptador de corriente ..................................................... 6
2.4 Conexión de los electrodos pH ................................................................. 7
2.5 Soporte para el pHmetro para ajustar el ángulo de visión ...................... 7
2.6 Guía rápida incluida .................................................................................... 8
3 Ajuste .................................................................................... 8
3.1 Ajustar la unidad de temperatura .............................................................. 8
3.2 Ajustar la compensación de la temperatura manual ............................... 8
3.3 Seleccionar un grupo predefinido de tampones ...................................... 8
4 Funcionamiento del STARTER 3100 .................................. 9
4.1 Calibración .................................................................................................. 9
4.1.1 Grupo de tampones ....................................................................... 9
4.1.2 mo realizar la calibración de un punto .................................... 10
4.1.3 mo realizar la calibración de dos puntos ................................. 11
4.2 Medición de la muestra ............................................................................ 11
4.2.1 Medición del pH ............................................................................ 11
4.2.2 Medición de mV ........................................................................... 12
4.3 Medición de la temperatura ...................................................................... 12
4.4 Uso de la memoria .................................................................................... 12
4.4.1 Almacenar una lectura ................................................................ 12
4.4.2 Recuperar desde la memoria ...................................................... 12
4.4.3 Borrar el contenido de la memoria .............................................. 12
4.5 Imprimir ..................................................................................................... 13
5 Mantenimiento ................................................................... 14
5.1 Mensajes de error ..................................................................................... 14
5.2 Mantenimiento del pHmetro ..................................................................... 14
5.3 Mantenimiento del electrodo ................................................................... 14
5.4 Autodiagnóstico ........................................................................................ 15
5.5 Recuperar los valores de fábrica ............................................................. 15
6 Datos técnicos ................................................................... 16
6.1 Especificaciones ....................................................................................... 16
7 Conformidad ...................................................................... 17
8 Grupo de tampones ........................................................... 18
STARTER 3100 ES-1
1 Introducción
Gracias por elegir un producto OHAUS.
Para evitar el funcionamiento incorrecto, lea el manual completamente antes de utilizar el
pHmetro de mesa STARTER 3100.
Los STARTER 3100 tienen una excelente relación calidad/precio y están diseñados con muchas
funciones útiles incluyendo un soporte para electrodo autónomo, una alarma para notificar un
error. También están disponibles en OHAUS otros accesorios como varios electrodos pH para
diferentes aplicaciones, y soluciones de tamn.
En estos pHmetros encontrará funcionalidades muy útiles. Algunas de las muchas funciones son:
Gran pantalla LCD retroiluminada con guía rápida incluida para ayudar al usuario a trabajar
con comodidad
Punto final automático y manual que puede ayudar a los usuarios a fijar los valores de
lectura estables
3 grupos de tampones predefinidos (EE.UU, UE y JJG) que cumplen con los diferentes
requisitos
RS232 puede conectar una impresora, memoria de 99 datos con los últimos datos de
calibración
1.1 Definición de palabras de advertencia y símbolos de advertencia
Las indicaciones de seguridad están marcadas con palabras de advertencia y símbolos de
advertencia. Muestran cuestiones y advertencias de seguridad. Hacer caso omiso de las
indicaciones de seguridad puede provocar lesiones personales, daños en el equipo, fallos de
funcionamiento y resultados falsos.
Palabras de advertencia
ADVERTENCIA Para una situación de peligro con riesgo medio, que podría
ocasionar lesiones o la muerte si no se evita.
PRECAUCIÓN Para una situación de peligro con riesgo alto, que podría dañar el
dispositivo o la propiedad o la rdida de datos, o lesiones si no se
evita.
Atención Para información importante sobre el producto.
Nota Para información útil sobre el producto
ES-2 STARTER 3100
mbolos de advertencia
Peligro general
Peligro de explosión
Peligro de corrosión
Corriente alterna
Corriente continua
1.2 Precauciones de seguridad
PRECAUCIÓN: Lea todas las advertencias de seguridad antes de la instalación, conexión o
reparación del equipo. No cumplir con estas advertencias podría causar daños corporales y/o
daños en la propiedad. Guarde las instrucciones para futuras consultas.
Compruebe que el rango de tensión de entrada impreso en la etiqueta de datos y el tipo de
enchufe coincidan con la alimentación por CA utilizada.
Asegúrese de que el cable de corriente no represente un posible obstáculo o peligro de
tropiezo.
Utilice el equipo solo en ambientes secos.
Seque inmediatamente cualquier líquido derramado. El instrumento no es hermético.
Cumpla las instrucciones del fabricante de productos químicos y las normas de seguridad
generales de laboratorio cuando utilice productos químicos y disolventes.
Utilice solo los accesorios y periféricos aprobados.
Utilice el equipo solo en las condiciones ambientales especificadas en estas instrucciones.
Desconecte el equipo de la fuente de alimentación cuando lo esté limpiando.
No utilice el equipo en entornos peligrosos o inestables.
El mantenimiento debe realizarse solamente por personal autorizado.
STARTER 3100 ES-3
1.3 Pantalla y controles
Pantallas
1
Condición del electrodo
Pendiente: más del 95 %
y compensación: ± (0-15) mV
La condición del electrodo es
buena
Pendiente: menos del 90 %
o compensación: ± (35-60) mV
La condición del electrodo no es
buena y necesita limpieza
2
Icono de estabilidad del punto final , icono de punto final automático
3
Icono de medición - , cuando parpadea, indica que la medición o la calibración se está
ejecutando
4
Icono de calibración - , cuando se muestra en la pantalla indica que la calibración está
en curso
5
Icono de ajuste - ; el instrumento está en modo ajuste, puede ajustar la temperatura
(MTC), el grupo de tampones, etc.
6
Lectura de pH/mV o la pendiente en el proceso de calibración
7
Punto de calibración / grupo de tampones /Número de memoria / Índice de error
8
Compensación automática de la temperatura - ATC, compensación manual de la
temperatura - MTC
9
Temperatura durante la medición o la compensación (mV) en el proceso de calibración
2
3
4
1
5
6
7
8
9
ES-4 STARTER 3100
Controles
Botón
Pulsar y soltar
Pulsar y mantener
durante 3 segundos
- Iniciar o finalizar la medición
- Confirmar ajuste, almacenar un
valor entero
- Encender/apagar punto final
automático
,
- Iniciar calibración
- Recuperar los últimos datos
de calibración: pendiente,
compensación
- Encender el pHmetro
Salir y volver a la pantalla de
medición
- Apagar el pHmetro
- Almacenar en memoria lecturas
actuales
- Aumentar el valor durante la
configuración
- Desplazarse hacia arriba por la
memoria
Recuperar datos almacenados
- Imprimir datos de memoria
actuales
- Cambiar entre pH y mV
- Disminuir el valor durante la
configuración
- Desplazarse hacia abajo por la
memoria
- Acceder al modo ajuste
- Iniciar autodiagnóstico
Encender/Apagar la
retroiluminación de la LCD
STARTER 3100 ES-5
2 Instalación
Desempaquete cuidadosamente la caja.
2.1 Contenido del paquete
El modelo ST3100-B (paquete básico) tiene los siguientes artículos:
Además del contenido del ST3100-B, el paquete de ST3100-F también incluye lo siguiente:
Además del contenido del ST3100-B, el paquete de ST3100-H también incluye lo siguiente:
Se debe disolver calda tampón de pH para polvo en 250 ml de agua pura o desionizada en un
matraz volumétrico. También puede solicitar el frasco con solución de tampón en Ohaus.
Ahora hay disponibles más electrodos que incluyen lo siguiente:
Modelo
Descripción
P/N
ST310
Electrodo de plástico para pH rellenable 3 en 1
83033965
ST210
Electrodo de plástico para pH rellenable 2 en 1
83033966
ST320
Electrodo de plástico para pH, con gel, 3 en 1 (no es
necesario rellenarlo)
83033967
ST230
Electrodo de vidrio para pH, muestras de lodo. 2 en 1
83033968
STORP1
Electrodo de plástico ORP, con gel
30038555
STORP2
Electrodo de vidrio ORP, rellenable
30038553
STTEMP30
Electrodo de temperatura
83033970
ST3100-B
Unidades
STARTER 3100
1
Soporte para electrodo autónomo
1 juego
Cubierta en uso
1
Fuente de alimentación de 12 V
1 juego
Juego de polvo para tampón de pH (4,01, 7,00, 10,01)
1 juego
Electrodo de pH rellenable 3 en 1 ST310
1
Juego de polvo para tampón de pH (4,01, 7,00, 10,01)
1 juego
Electrodo de pH gel 3 en 1 ST320
1
ES-6 STARTER 3100
Tampones y accesorios:
Juego de tampón para polvo (4,01; 7,00; 10,00)
83033971
Tampón con pH 1,68, 250 ml
30100424
Tampón con pH 4,01, 250 ml
30100425
Tampón con pH 7,00, 250 ml
30100427
Tampón con pH 10,01, 250 ml
30100429
Tampón con pH 12,45, 250 ml
30100430
Soporte para electrodo autónomo (3100/3100C)
30058733
Cubierta en uso
30058734
2.2 Instalación del soporte para electrodo autónomo
Coloque el brazo del electrodo a la base,
Después de ajustar hasta cierto punto el mando de tensión, puede mover hacia arriba y
hacia abajo la parte superior del brazo directamente. Al colocar el electrodo del pH en el
brazo y realizar la calibración o medición, se puede colocar el frasco de almacenamiento del
electrodo de pH en el hueco de la base.
2.3 Instalación del adaptador de corriente
Inserte el clip adaptador correcto en la ranura del adaptador de corriente. La fuente de alimentación de
la balanza Ohaus es también de 12 V que puede utilizarse tambn para el pHmetro.
STARTER 3100 ES-7
2.4 Conexión de los electrodos pH
Existen 3 enchufes por electrodo: el enchufe "pH" (BNC), “TEMP”, el enchufe (Cinch) y el enchufe
“REF” (banana de 2 mm).
Para el electrodo ST310 debe conectar el enchufe BNC y Cinch. Para el electrodo de pH 2 en 1
(p. ej. ST210) solo necesita conectar BNC.
El enchufe REF es para el electrodo separado de referencia. (p. ej. STREF01)
2.5 Soporte para el pHmetro para ajustar el ángulo de visión
Un diseño único del STARTER 3100 es el soporte para el pHmetro, que puede utilizarse para
ajustar el ángulo de visión de la pantalla en caso de trabajar en mesas de laboratorio altas.
ES-8 STARTER 3100
2.6 Guía rápida incluida
Otro diseño único del STARTER 3100 es la guía rápida incluida. La guía rápida se incluye en el
interior de la parte inferior de la carcasa del pHmetro, puede sacarla con los dedos.
3 Ajuste
3.1 Ajustar la unidad de temperatura
Tenga en cuenta:
Si se utiliza un electrodo de temperatura, compensación automática de temperatura (ATC) y la
temperatura de la muestra se muestran en la pantalla. A continuación, puede optar por omitir la
configuración MTC (abajo).
Si el medidor no detecta un electrodo de temperatura o no se utiliza, el medidor cambia
automáticamente al modo de compensación de temperatura manual (MTC) y MTC aparece en la
pantalla.
MTC se puede ajustar de la siguiente manera:
Poder del medidor pulsando Exit (Salir).
Mantenga pulsado el botón Mode/Setup (Modo/Ajuste) hasta que aparezca el icono en
la pantalla y el estándar actual parpadee.
Pulse los botones Up (Arriba) o Down (Abajo) para cambiar entre °C y °F.
Pulse el botón Read (Leer) para confirmar su selección.
Después
Continúe con la configuración de la temperatura MTC o
Pulse el botón Read (Leer) para confirmar su selección.
Pulse el botón Exit (Salir) volver a la pantalla de medición.
3.2 Ajustar la compensación de la temperatura manual
Después de confirmar la unidad de la temperatura, se mostrará en la pantalla el valor de la
temperatura MTC (Compensación de la temperatura manual). Después utilice los botones up
(arriba) o down (abajo) para aumentar o disminuir el valor de la temperatura de su muestra.
Pulse el botón Read (Leer) para confirmar el ajuste o pulse el botón Exit (Salir) para volver a la
pantalla de medición.
El ajuste del valor de la temperatura MTC por defecto es 25 °C (77 °F).
Nota: °C = 5/9 (°F -32)
3.3 Seleccionar un grupo predefinido de tampones
Después de confirmar el valor de la temperatura MTC de compensación, va hasta la selección del
grupo de tampones. Utilice los botones Up (Arriba) o Down (Abajo) para seleccionar un grupo de
tampones entre los 4 grupos de tampones (3.1.1). Pulse el botón Read (Leer) para confirmar el
ajuste o pulse el botón Exit (Salir) para volver a la pantalla de medición.
El grupo de tampones por defecto es b1.
b1
1,68
4,01
7,00
10,01
STARTER 3100 ES-9
4 Funcionamiento del STARTER 3100
El procedimiento de la medicn del pH es el siguiente, los puntos s importantes son el
electrodo de pH, aclarado, calibración y almacenamiento. Después de colocar el electrodo en la
solución, primero debemos mezclarlo y después realizar la operación.
a) Preparar el electrodo de pH
b) Preparar el tampón y la calibración del electrodo de pH
c) Preparar la muestra
d) Medir el pH
e) Registrar o imprimir del resultado de la medición
f) Enjuagar y almacenar el electrodo de pH
Preparación del electrodo de pH: el electrodo de pH debe enjuagarse con agua pura antes y
después de su uso. Compruebe si el electrodo está sicamente dañado (tenga cuidado con la
pera de vidrio)
El electrodo de pH debe almacenarse en el frasco de almacenamiento, la solución del frasco es
una solución de KCl 3M. Después de colocar el electrodo de pH en la solución de la muestra o
del tampón, el usuario debe mezclar durante varios segundos, después esperar de 30 a 60
segundos para que la señal se estabilice, y después pulsar el botón para comenzar a trabajar
(calibración o medición).
ADVERTENCIA No use el equipo en ambientes peligrosos. El equipo no está
protegido contra explosiones.
ADVERTENCIA Cumpla las instrucciones del fabricante de productos químicos y las
normas de seguridad generales de laboratorio cuando utilice productos químicos y
disolventes.
4.1 Calibración
4.1.1 Grupo de tampones
El STARTER 3100 puede realizar calibraciones de 1-, 2- o 3- puntos.
En el pHmetro hay 3 grupos de tampones, puede seleccionar el grupo de tampones que prefiera
(consulte 4.3), el grupo de tampones por defecto es el estándar EE.UU b1 y durante la
calibración se reconocerá automáticamente el valor del tampón. Los 3 grupos predefinidos de
tampones son (a 25 °C):
El STARTER 3100 corrige automáticamente la dependencia de la temperatura de los valores de
pH del tampón que figuran en la siguiente tabla, grupo del tampón b1.
Esto quiere decir, que si la temperatura de la solución del tampón es de 15 °C, el valor de la
calibración que obtendrá debería ser pH 1,67 (tampón 1,68), pH 4,00 (tampón 4,01), pH 7,04
(tampón 7,00) y pH 10,12 (tampón 10,01).
b1
1,68
4,01
7,00
10,01
b2
2,00
4,01
7,00
9,21
11,00
b3
1,68
4,00
6,86
9,18
12,46
ES-10 STARTER 3100
Calibración: Es necesario calibrar el electrodo de pH con la solución de tampón esndar de pH
antes de realizar una medición de pH adecuada. La calibración sirve para mostrar el valor de pH
correcto cuando el pHmetro recibe la señal del valor mV del electrodo de pH.
Pendiente: el coeficiente lineal entre mV y pH según el valor teórico (p.ej. -59,16 mV/pH
@ 25 °C lo que indica una pendiente del 100 %);
Compensación: el valor de mV cuando el valor del pH es 7,00. (El valor teórico es 0
mV);
4.1.2 mo realizar la calibración de un punto
Al realizar la calibración, Ohaus recomienda utilizar el modo de punto Colgar automáticamente.
Después de encender el medidor, asegúrese de que la parte superior de la pantalla muestra
para asegurar que el medidor está en modo de punto auto End.
Modo de punto final: Hay dos modos de punto final, punto final automático y punto final
manual. Al mantener pulsado el botón Read (Leer), puede cambiar entre los modos de punto
final.
o Para llevar al punto final manualmente una medición o calibración, tiene que
pulsar el botón Read (Leer): La lectura de la muestra queda fija, el icono de
estabilidad del punto final parpadeará 3 veces y se quedará fijo en la pantalla.
El pHmetro no puede alcanzar automáticamente el punto final.
o Cuando se encuentre en el modo de punto final automático, el pHmetro decide si
la lectura es estable y después llega automáticamente al punto final, la lectura
queda fija y parpadeará 3 veces y después desaparecerá, los iconos de punto
final automático parpadearán 3 veces y quedarán fijos en la pantalla.
o Criterio de estabilidad: La señal de la entrada del sensor puede no variar en más
de 0,1 mV en 6 segundos.
Nota: con la calibración de un punto solo se ajusta la compensación. Si el sensor se calibró
previamente con calibración multipunto, se mantendrá la pendiente memorizada previamente. Si
no, se utilizará la pendiente del 100% teórica (-59,16 mV / pH).
5 °C
1,67
4,01
7,09
10,25
10 °C
1,67
4,00
7,06
10,18
15 °C
1,67
4,00
7,04
10,12
20 °C
1,68
4,00
7,02
10,06
25 °C
1,68
4,01
7,00
10,01
30 °C
1,68
4,01
6,99
9,97
35 °C
1,69
4,02
6,98
9,93
40 °C
1,69
4,03
6,97
9,89
45 °C
1,70
4,05
6,97
9,86
50 °C
1,71
4,06
6,96
9,83
STARTER 3100 ES-11
Cuando el STARTER 3100 se encuentre en modo medición del pH (consulte 3.2), coloque el
electrodo de pH en un tampón de calibración, mezcle durante 5 segundos, espere 30 segundos y
después:
Pulse el bon Cal. (Calibración), se mostrará Cal 1 en la parte inferior izquierda de la
pantalla y “Cal ” parpadeará. y aparecerán en la parte superior de la pantalla,
parpadeará durante la calibración.
El pHmetro alcanzará el punto final según el modo de punto final, se mostrará en la
pantalla el valor del pH del punto de calibración (p. ej. 7,00) junto con la temperatura.
Ha finalizado la calibración de 1 punto, ahora existen 3 opciones para realizar a continuación:
Pulse el botón Cal (Calibración) para realizar la calibración de 2 puntos.
Pulse el botón Read (Leer) para almacenar la calibración de 1 punto y salir, durante 3
segundos se mostrarán en la pantalla la compensación y la pendiente y después volverá
a la pantalla de medición.
Pulse el botón Exit (Salir) para rechazar la calibración, vuelva a la pantalla de medición.
4.1.3 mo realizar la calibración de dos puntos
Realizar la calibración de un punto como se describe anteriormente.
Enjuagar el electrodo de pH con agua pura y secar con un pañuelo de papel.
Colocar el electrodo en el siguiente tampón de la calibración, mezclar y esperar, después
pulsar el botón Cal (Calibración), aparecerá Cal 2y Cal ”parpadeará. En la parte superior
de la pantalla aparecerá y , durante la calibración parpadeará .
El pHmetro alcanzará el punto final según el modo de punto final, se mostrará en la
pantalla el valor del pH del punto de calibración (p. ej. 4,01) junto con la temperatura.
Ha finalizado la calibración de dos puntos, ahora existen tres opciones para realizar a
continuación:
Pulse el botón Cal (Calibración) para realizar la calibración de tres puntos.
Pulse el botón Read (Leer) para almacenar la calibración de dos puntos y salir, durante tres
segundos se mostrarán en la pantalla la compensación y la pendiente y después volverá
a la pantalla de medición.
Pulse el botón Exit (Salir) para rechazar la calibración, vuelva a la pantalla de medicn.
Nota: se recomienda el uso de un electrodo de temperatura o un electrodo con un sensor de
temperatura integrado (electrodo de pH 3 en 1). Si utiliza el modo MTC, debe introducir el valor
de temperatura correcto y mantener todas las soluciones de los tampones y de las muestras a la
temperatura establecida. (consulte 4.2)
4.2 Medición de la muestra
4.2.1 Medición del pH
Coloque el electrodo de pH en la muestra. Enjuague durante cinco segundos, espere 30
segundos.
Pulse el botón Read (Leer) para comenzar la medición del pH. aparecerá en la pantalla .
Durante la medición parpadeará.
Cuando el pHmetro alcance el punto final, se mostrará en la pantalla el valor del pH junto
con la temperatura.
ES-12 STARTER 3100
4.2.2 Medición de mV
Pulse el botón Mode (Modo) para cambiar entre el modo de medición del pH y el modo
de medición de mV.
Siga el mismo procedimiento de medición del pH para realizar una medición de mV.
4.3 Medición de la temperatura
Para mayor precisión, recomendamos el uso de un electrodo de temperatura integrado o
independiente.
Si se utiliza un electrodo de temperatura, se mostrarán en la pantalla la ATC y la
temperatura de la muestra.
Si el pH metro no detecta un electrodo de temperatura, cambiará automáticamente al modo
de compensación de temperatura manual y aparecerá MTC. Debe fijarse la temperatura
MTC.
Nota: el STARTER 3100 acepta el sensor de temperatura NTC 30 . ATC (compensación
automática de la temperatura) o MTC (compensación manual de la temperatura), la
compensación de la temperatura solo se corrige por el cambio de la salida del electrodo, no por el
cambio en la solución actual. Esto quiere decir que el pHmetro corrige la señal (mV) de los
electrodos pH para obtener una mayor exactitud del valor del pH según la temperatura real.
4.4 Uso de la memoria
4.4.1 Almacenar una lectura
El STARTER 3100 puede almacenar hasta 99 resultados de punto final.
Pulse el botón Store (Almacenar) cuando la medición haya llegado al punto final. M01
indica que se ha almacenado un resultado.
Si pulsa el botón Store (Almacenar) cuando se muestre M99, se mostrará FUL (Lleno) para
indicar que no hay s espacio libre en la memoria. Para almacenar s datos deberá borrar el
contenido de la memoria. (consulte 3.4.3).
4.4.2 Recuperar desde la memoria
Cuando las mediciones lleguen al punto final, mantenga pulsado el botón Recall (Recuperar)
para recuperar los valores de la memoria almacenados.
Pulse los botones up (arriba) o down (abajo) para desplazarse por los resultados
almacenados. De R01 a R99 indica qué resultado se está visualizando.
Pulse el botón Exit (Salir) para salir.
4.4.3 Borrar el contenido de la memoria
Pulse los botones up (arriba) o down (abajo) para desplazarse por los resultados
almacenados hasta que aparezca “MRCL” (Borrar memoria).
Pulse el botón Read (Leer), CLr (Borrar) parpadeará.
Ahora tenemos dos opciones:
Pulsar el botón Read (Leer) para confirmar la eliminacn de todos los datos almacenados.
Pulsar el botón Exit (Salir) para volver al modo de medición sin eliminar la memoria.
STARTER 3100 ES-13
4.5 Imprimir
Si una impresora es conectada al STARTER 3100 (p.ej. SF-F40A, la luz verde indica que la
conexión se ha realizado con éxito, la tasa de baudios de la impresora es de 9600 bps, 8 bits de
datos, sin bit de paridad, 1 bit de parada), se generará automáticamente una impresión después
de que cada medición o calibración alcance el punto final.
El formato para la impresión después de una medición de pH es:
Punto final, valor, Temp., ATC/MTC
EP automático, 4,01 pH, 25,0 °C, MTC
Los detalles para la segunda línea son:
EP automático, 4,01 pH, 25,0 °C, MTC
| | | |---- compensación manual de la temperatura
| | |----------------Valor de la temperatura y unidad °C
| |------------------------- valor del pH
|---------------------------------------- Punto final automático
El formato para la impresión después de una medición de mV es:
Punto final, valor, Temp., ATC/MTC
EP manual, 182 mV, 23,2 °C , ATC
La impresión en caso de un mensaje de error es:
Punto final, valor, Temp., ATC/MTC
Error3
La impresión para una calibración de 2 puntos con éxito es:
Tampón1: 4,01 pH
mV1: 178 mV
Temp.1: 25,0 °C
Tampón2: 7,00 pH
mV2: 3 mV
Temp.2: 25,0 °C
Pendiente: 99 %
Compensación: 5 mV
ATC/MTC: MTC
La impresión para una calibración de 3 puntos con éxito es:
Tampón1: 4,01 pH
mV1: 178 mV
Temp.1: 25,0 °C
Tampón2: 7,00 pH
mV2: 0 mV
Temp.2: 25,0 °C
Tampón3: 9,21 pH
mV3: -130 mV
Temp.3: 25,0 °C
Pendiente: 100%
Compensación: 0 mV
ATC/MTC: ATC
Imprimir desde la memoria. Al desplazarse por la memoria puede imprimir la entrada que se está
visualizando manteniendo pulsado el botón Store (Almacenar) durante 3 segundos. El formato de
impresión es el siguiente:
Recuperar 01:
Punto final, valor, Temp., ATC/MTC
EP automático, 4,01 pH, 35 °C, ATC
ES-14 STARTER 3100
5 Mantenimiento
5.1 Mensajes de error
Error 0
Error de acceso a la memoria
Restablezca a los valores de fábrica
Error 1
Fallo del autodiagnóstico
Repita el procedimiento del autodiagnóstico y
asegúrese de que finaliza pulsando las cinco
teclas en el espacio de dos minutos.
Error 2
Valores de medición fuera del
intervalo
Compruebe que el electrodo está correctamente
conectado y colocado en la solucn de muestra.
Error 3
Temperatura de medición del
tampón fuera del intervalo
(<5 o >40 °C)
Mantenga la temperatura del tamn del pH en el
intervalo de calibración
Error 4
Compensación fuera del
intervalo
Compensación > 60 mV o < -
60 mV
Asegúrese de que el tampón pH es el correcto y
es reciente. Limpie o reemplace el electrodo de
pH.
Error 5
Pendiente fuera del intervalo
Asegúrese de que el tampón es el correcto y es
reciente. Limpie o reemplace el electrodo de pH.
Error 6
El pHmetro no reconoce el
tampón
Asegúrese de que el tampón es correcto y es
reciente, compruebe que el tampón no se ha
utilizado más de una vez.
Error 9
El conjunto de datos actuales
ya se ha almacenado una
vez
Una lectura de punto final solo puede
almacenarse una vez. Realice una nueva
medición para almacenar.
Error 10
La temperatura de muestra
se encuentra fuera del
intervalo
Compruebe la temperatura de muestra, el sensor
de temperatura
Además, si ocurre un error, el pHmetro sonará tres veces para avisar.
Para cualquier otro problema, póngase en contacto con el representante de ventas de Ohaus.
5.2 Mantenimiento del pHmetro
¡No desatornille nunca las dos mitades de la carcasa!
Los equipos de la serie STARTER 3100 no necesitan ningún otro mantenimiento que una
limpieza ocasional con un paño húmedo y el reemplazo de pilas gastadas.
La carcasa está fabricada en ABS (acrilonitrilo-butadieno-estireno). Este material es sensible a
algunos disolventes orgánicos como tolueno, xileno y metiletilcetona (MEK). Debe limpiarse
inmediatamente cualquier derrame.
5.3 Mantenimiento del electrodo
Asegúrese de que el electrodo esté lleno de solución electrolítica. Almacene siempre el electrodo
según indique el manual de instrucciones del electrodo y no permita que se seque.
STARTER 3100 ES-15
Si se ralentiza la respuesta del electrodo o la pendiente no es suficientemente buena, intente:
o Sumergir el electrodo en 0,1 M HCI durante más de 8 horas.
o Para los contaminantes de grasas y aceites, desengrase la membrana con un algodón
empapado en acetona o una solución de jabón.
Después del tratamiento del electrodo, debe realizarse una nueva calibración. Si la pendiente del
electrodo sigue sin ser correcta, puede necesitar cambiar el electrodo.
5.4 Autodiagnóstico
Cuando el pHmetro esté encendido, mantenga pulsado el botón Read (Leer) y el botón Cal
(Calibración) simultáneamente hasta que el pHmetro muestre la pantalla completa. Los
iconos parpadearán uno tras otro.
De esta manera podrá comprobar si se muestran correctamente todos los iconos. El siguiente
paso es el de comprobar que las teclas funcionan correctamente. Para esto es necesaria la
interacción del usuario.
Cuando b parpadee, se mostrarán cinco iconos.
Pulse las cinco teclas en cualquier orden. Cada vez que pulse una tecla, desaparecerá un
icono de la pantalla, continúe pulsando las demás teclas hasta que hayan desaparecido
todos los iconos.
Cuando haya finalizado el autodiagnóstico con éxito, aparecerá PAS (Superado). Si el
autodiagnóstico falla, aparecerá el mensaje de error Err 1 (Error 1).
Nota: Debe finalizar pulsando las cinco teclas en el espacio de 2 minutos, de otra forma
aparecerá el mensaje Err 1 (Error 1) y debe repetir el procedimiento.
5.5 Recuperar los valores de fábrica
Cuando el pHmetro esté apagado, mantenga pulsados a la vez durante 3 segundos los
botones Read (Leer), Cal (Calibración) y Exit (Salir), la pantalla mostrará y
parpadeará, lo que indicará que está "reiniciando". Tendremos entonces dos opciones:
Pulsar el botón Read (Leer) para restablecer los valores de brica (MTC, pendiente,
compensación, etc.), se mostrará después reinicie el pHmetro.
O pulsar el botón Exit (Salir) para salir de la configuración, se mostrará después
apague el pHmetro.
ES-16 STARTER 3100
6 Datos técnicos
6.1 Especificaciones
Condiciones ambientales
Solo para uso en interiores
Altitud: Hasta 2000 m
Intervalo de temperatura especificado: 5 °C a 40 °C
Humedad: humedad xima relativa del 80 % para temperaturas de hasta 30 °C reducción
de linealidad hasta una humedad relativa del 50 % a 40 °C
Fluctuaciones en el voltaje de la alimentación eléctrica: hasta ±10% del voltaje nominal
Categoría de instalación II
Grado de polución: 2
El funcionamiento se asegura a temperaturas ambientales de entre 5 °C a 40 °C
Modelo
ST3100
Intervalo de medición
-2,00…16,00 pH
–1999…1999 mV
-5 °C …110 °C
Resolución
0,01 pH
1 mV
0,1 °C
mites de error
± 0,01 pH
± 1 mV
± 0,5 °C
Calibración
3 puntos
3 grupos predefinidos de tampones
Memoria
99 mediciones
El último dato de calibración
Suministro de corriente
Entrada para adaptador de CA: 100-240 V ~ X.XA 50/60
Hz
Salida para adaptador de CA: 12 V X.XA
Tamaño/peso
220 (An) x 175 (Pr) x 78 (Al) (mm) / 0,75 Kg
Pantalla
Cristal líquido con retroiluminación
Entrada
BNC, impedancia > 10e+12 Ω
Cinch, NTC 30 kΩ
Entrada de referencia
Enchufe tipo banana de 2 mm
Compensación de temperatura
ATC y MTC
Carcasa
ABS
STARTER 3100 ES-17
7 Conformidad
Este producto cumple con la directiva EMC 2004/108/CE y con la directiva de
baja tensión 2006/95/CE. La declaración de conformidad está disponible a
través de Internet en europe.ohaus.com/europe/en/home/support/weee.aspx.
En cumplimiento con la Directiva europea 2002/96/CE sobre residuos de
aparatos eléctricos y electrónicos (WEEE), este dispositivo no se puede
eliminar como residuo doméstico. También se aplica en países fuera de la UE,
según sus requisitos específicos. Elimine este producto según las normativas
locales en el punto de recolección especificado para equipos eléctricos y
electrónicos. Si tiene alguna pregunta, póngase en contacto con la autoridad
responsable o el distribuidor del que adquirió este dispositivo. En caso de que
este dispositivo se entregará a un tercero (para uso privado o profesional), se
debe hacer referencia al contenido de esta regulación. Las instrucciones de
eliminación para Europa están disponibles a través de Internet en
europe.ohaus.com/europe/en/home/support/weee.aspx. Gracias por su
contribución a la protección del medioambiente.
Nota sobre la FCC: Este equipo ha sido sometido a pruebas y ha demostrado cumplir con
los límites para dispositivos digitales de clase A, según la Parte 15 de las normas de la FCC.
Estos límites están diseñados para proporcionar una protección razonable frente a
interferencias perjudiciales cuando se utiliza el equipo en un entorno comercial. Este equipo
genera, utiliza y puede emitir energía de radiofrecuencia y, si no se instala y utiliza según
las instrucciones del manual puede provocar interferencias perjudiciales a las
comunicaciones por radio. La utilización de este equipo en un área residencial puede
provocar interferencias perjudiciales. En este caso, el usuario deberá corregir la
interferencia por su cuenta.
Registro en la ISO 9001
En 1994, OHAUS Corporation, EEUU, recibió el certificado de registro en la ISO 9001 del
Bureau Veritus Quality International (BVQI), confirmando que el sistema de gestión de
calidad de OHAUS cumple con los requisitos de la norma ISO 9001. El 21 de junio de 2012,
OHAUS Corporation, EEUU, se registró nuevamente en la norma ISO 9001:2008.
ES-18 STARTER 3100
8 Grupo de tampones
El STARTER 3100 corrige automáticamente la dependencia de la temperatura del valor de pH del
grupo del tampón que se facilita en las siguientes tablas (b2, b3), puede encontrar b1 (consulte
3.1.1)
Grupo de tampón b2 Estándar europeo
Temp °C pH 2,00 pH 4,01 pH 7,00 pH 9,21 pH 11,00
5 2,02 4,01 7,09 9,45 11,72
10 2,01 4,00 7,06 9,38 11,54
15 2,00 4,00 7,04 9,32 11,36
20 2,00 4,00 7,02 9,26 11,18
25 2,00 4,01 7,00 9,21 11,00
30 1,99 4,01 6,99 9,16 10,82
35 1,99 4,02 6,98 9,11 10,64
40 1,98 4,03 6,97 9,06 10,46
45 1,98 4,04 6,97 9,03 10,28
50 1,98 4,06 6,97 8,99 10,10
Grupo de tampón b3 JJG119
Temp °C pH 1,68 pH 4,00 pH 6,86 pH 9,18 pH 12,46
5 1,67 4,00 6,95 9,39 13,21
10 1,67 4,00 6,92 9,33 13,01
15 1,67 4,00 6,90 9,28 12,82
20 1,68 4,00 6,88 9,23 12,64
25 1,68 4,00 6,86 9,18 12,46
30 1,68 4,01 6,85 9,14 12,29
35 1,69 4,02 6,84 9,11 12,13
40 1,69 4,03 6,84 9,07 11,98
45 1,70 4,04 6,83 9,04 11,83
50 1,71 4,06 6,83 9,02 11,70
STARTER 3100 ES-19
GARANTÍA LIMITADA
Los productos de Ohaus están en garantía por defectos en los materiales o de construcción
desde la fecha de entrega hasta la duración del periodo de la garantía. Durante el periodo de
validez de la garantía, Ohaus reparará o, según su criterio, sustituirá cualquier componente que
sea defectuoso sin cargo adicional siempre que el producto se entregue, con flete pago, a Ohaus.
Esta garantía no aplica si el producto ha recibido daños por accidentes o mal uso, ha sido
expuesto a materiales radioactivos o corrosivos, ha recibido materiales extraños en su interior o
como resultado del servicio o modificación por un tercero que no sea Ohaus. En lugar de una
tarjeta de registro de garantía debidamente devuelta, el periodo de la garantía se iniciará en la
fecha de envío al distribuidor autorizado. Ohaus Corporation no entrega otro tipo de garantía
expresa o implícita. Ohaus Corporation no es responsable de daños indirectos.
Puesto que la legislación sobre garantías difiere de un estado a otro y de un país a otro, póngase
en contacto con Ohaus o con su distribuidor local de Ohaus para obtener más información.
Table des matières
1 Introduction .......................................................................... 1
1.1 finition des symboles et des indicateurs d'avertissements ... 1
1.2 Précautions de sécurité .................................................................. 2
1.3 Contrôles et affichage .................................................................... 3
2 Installation ............................................................................ 5
2.1 Contenu de l’emballage .................................................................. 5
2.2 Installation du porte-électrode autonome .................................... 6
2.3 Installation de l'adaptateur d'alimentation .................................... 6
2.4 Connexion des électrodes de pH .................................................. 7
2.5 Support de réglage de l'angle de vue de l'appareil de mesure .. 7
2.6 Guide rapide attaché ....................................................................... 7
3 Configuration ....................................................................... 8
3.1 finition de l'unité de température .............................................. 8
3.2 finition de la compensation manuelle de la température ....... 8
3.3 lection d'un groupe de tampons préfinis ............................. 8
4 Fonctionnement du STARTER 3100 ................................... 9
4.1 Étalonnage ....................................................................................... 9
4.1.1 Groupe de tampons ....................................................................... 9
4.1.2 alisation d'un étalonnage à 1 point ......................................... 10
4.1.3 alisation d'un étalonnage à 2 points ........................................ 11
4.2 Mesure d'un échantillon ............................................................... 11
4.2.1 Mesure du pH ............................................................................... 11
4.2.2 Mesure mV .................................................................................. 12
4.3 Mesure de température ................................................................. 12
4.4 Utilisation de la mémoire .............................................................. 12
4.4.1 Stockage d'une mesure ............................................................... 12
4.4.2 Rappel depuis la mémoire .......................................................... 12
4.4.3 Effacement de la mémoire .......................................................... 12
4.5 Impression ..................................................................................... 13
5 Maintenance ....................................................................... 14
5.1 Message d’erreur ........................................................................... 14
5.2 Maintenance de l'appareil de mesure .......................................... 14
5.3 Maintenance de l'électrode .......................................................... 14
5.4 Auto-diagnostic ............................................................................. 15
5.5 cupération des paramètres usine ........................................... 15
6 DONNÉES TECHNIQUES .................................................. 16
6.1 Spécifications ................................................................................ 16
7 Conformité.......................................................................... 17
8 Groupe de tampons ........................................................... 18
STARTER 3100 FR-1
1 Introduction
Merci d'avoir choisi un produit OHAUS.
Lire le manuel complètement avant d'utiliser le pH-mètre de laboratoire STARTER 3100 afin d'éviter
une exploitation inadéquate.
Le STARTER 3100 posde un rapport performances/prix excellent et est conçu avec de
nombreuses fonctions utiles, notamment un support d'électrode autonome et un bipeur pour les
alertes d'erreurs. D'autres accessoires comme de nombreuses électrodes de pH pour différentes
applications, et des solutions tampons en bouteilles sont aussi disponibles.
Vous trouverez de nombreuses fonctionnalis utiles sur ces appareils de mesure. En voici
quelques-unes ici présentées :
Grand ACL à troéclairage avec un Guide rapide attaché facilite l'exploitation par l'utilisateur
sultat final automatique/manuel qui peut aider les utilisateurs à figer des mesures stables
3 groupes de tampons préfinistats-Unis, EU et JJG) sont conformes à différentes
spécifications
RS232 peut connecter une imprimante, 99 mémoires de données avec les dernières
données d'étalonnage
1.1 finition des symboles et des indicateurs d'avertissements
Les remarques de curité sont marquées par des mots indicateurs et par des symboles
d'avertissement. Elles présentent les problèmes et les avertissements de curité. Ne pas
respecter les remarques de sécurité peut conduire à des accidents, à l'endommagement de
l'instrument, à des dysfonctionnements et à des résultats erronés.
Mots indicateurs
AVERTISSEMENT Pour une situation dangereuse avec un risque moyen pouvant être
à l'origine d'un accident ou d'un dés, s'il n'est pas évité.
PRÉCAUTIONS Pour une situation dangereuse avec un faible risque pouvant être à
l'origine de dommages au dispositif ou aux biens, à une perte des
données, ou à un accident, s'il n'est pas évité.
Attention Pour des informations importantes sur ce produit
Remarque Pour des informations utiles sur le produit
FR-2 STARTER 3100
Symboles d'avertissement
Danger !
Risque d'explosion
Danger, produits corrosifs
Tension alternative
Courant continu
1.2 Précautions de sécurité
ATTENTION : Lire tous les avertissements de sécurité avant d’effectuer l’installation, les
connexions ou la maintenance de cet équipement. Si ces avertissements ne sont pas respectés,
des accidents et/ou des dommages matériels peuvent en résulter. Conserver toutes les
instructions pour rérence ultérieure.
rifier que la plage de tension d'entrée imprimée sur l'étiquette des données et le type de
connecteur correspondent à l'alimentation CA locale à utiliser.
S'assurer que le cordon d'alimentation n’est pas un obstacle potentiel ou ne présente aucun
danger de chute.
Utiliser l'équipement seulement dans des lieux secs.
cher imdiatement tout déversement de liquide. L'instrument n'est pas étanche.
Avec des produits chimiques et des solvants, se conformer aux instructions du fabricant des
produits chimiques et aux règles gérales de sécurité du laboratoire.
Utiliser uniquement les accessoires et les périphériques approuvés.
Faire fonctionner l'équipement uniquement dans les conditions ambiantes spécifiées dans ces
instructions.
connecter l'équipement de l'alimentation électrique pour le nettoyer.
Ne pas utiliser l'équipement dans des environnements dangereux ou instables.
La maintenance doit être exclusivement exécutée par un personnel autorisé.
STARTER 3100 FR-3
1.3 Contrôles et affichage
Affichages
1
Condition de l'électrode
Pente : plus de 95 %
et décalage : ± (0-15) mV
Bonne condition de
l'électrode
Pente : moins de 90 %
ou décalage : ± (35-60) mV
La condition de l'électrode
n'est pas bonne ou
cessite un nettoyage
2
Icône de stabilité du résultat final ; icône du résultat final automatique
3
Icône de mesure - ; signifie mesure ou étalonnage en cours lorsqu'elle clignote
4
Icône d'étalonnage - ; signifie un étalonnage en cours lorsqu'elle clignote
5
Icône de configuration - ; instrument en mode de configuration, possibilité de finir la
température(MTC), le groupe de tampons, etc.
6
Lecture du pH/mV ou pente du processus d'étalonnage
7
Point d'étalonnage / Groupe tampon /Numéro mémoire / Index d'erreur
8
Compensation automatique de température : ATC ; compensation manuelle de température : MTC
9
Température pendant la mesure ou décalage (mV) du processus d'étalonnage
2
3
4
1
5
6
7
8
9
FR-4 STARTER 3100
Contrôles
Bouton
Appuyer et relâcher
Appuyer et maintenir
enfoncé
pendant 3 secondes
- Démarrer ou terminer une
mesure
- Confirmation du réglage,
stockage de la valeur entrée
- Marche/arrêt du résultat final
automatique
,
- Démarrage de l'étalonnage
- Rappel des dernières
données d'étalonnage :
pente et décalage
- Appareil de mesure en marche
- Quitter et revenir à l'écran de
mesure
- Appareil de mesure à l'arrêt
- Stockage de la mesure en cours
vers la mémoire
- Augmentation de la valeur
pendant les réglages
- Défilement vers le haut de la
moire
- Rappel des données
stockées
- Impression des données
moire en cours
- Commutation entre pH et mV
- Diminution de la valeur pendant
les réglages
- Défilement vers le bas de la
moire
- Entrer dans le mode de
configuration
- Démarrage des auto-diagnostics
Marche/arrêt du troéclairage
de l'ACL
STARTER 3100 FR-5
2 Installation
baller soigneusement la boîte.
2.1 Contenu de l’emballage
Le modèle ST3100-B (conditionnement de base) contient les articles suivants :
En supplément au contenu du ST3100-B, le conditionnement du ST3100-F comprend aussi ce qui suit :
En supplément au contenu du ST3100-B, le conditionnement du ST3100-H comprend aussi ce qui suit :
La poudre pour tampon de pH doit être dissoute dans 250 ml d'eau pure ou sionisée dans une
fiole jaue. Vous pouvez aussi commander auprès d'Ohaus des solutions de tampons en bouteille.
Un plus grand nombre d'électrodes est désormais disponible et comprend ce qui suit :
Modèle
Description
P/N
ST310
Électrode de pH réutilisable en plastique 3 en 1
83033965
ST210
Électrode de pH à remplissage en plastique 2 en 1
83033966
ST320
Électrode de pH sous forme de gel en plastique 3 en 1 (nul
besoin de remplissage)
83033967
ST230
Électrode de pH en verre pour échantillons troubles 2 en 1
83033968
STORP1
Électrode ORP en plastique pour gel
30038555
STORP2
Électrode réutilisable ORP en verre
30038553
STTEMP30
Électrode de température
83033970
ST3100-B
Unités
STARTER 3100
1
Porte-électrode autonome
1 ensemble
Couvercle en utilisation
1
Alimentation 12 volts
1 ensemble
Poudre pour tampon de pH (4,01, 7,00, 10,01)
1 ensemble
Électrode de pH réutilisable ST310 3 en 1
1
Poudre pour tampon de pH (4,01, 7,00, 10,01)
1 ensemble
Électrode de pH gel ST320 3 en 1
1
FR-6 STARTER 3100
Pièces et accessoires :
Poudre pour tampon de pH (4,01 ; 7,00 ; 10,00)
83033971
Tampon pH 1,68 250 ml
30100424
Tampon pH 4,01 250ml
30100425
Tampon pH 7,00 250ml
30100427
Tampon pH 10,01 250ml
30100429
Tampon pH 12,45 250ml
30100430
Porte-électrode autonome (3100/3100C)
30058733
Couvercle en utilisation
30058734
2.2 Installation du porte-électrode autonome
Installer le bras de l'électrode sur la base,
Après avoir glé le bouton de tension jusqu'à un certain niveau, il est possible de placer
directement le bras supérieur vers le haut et vers le bas. En plaçant l'électrode de pH sur le
bras et en alisant un étalonnage ou une mesure, il est possible de ranger la bouteille de
stockage de l'électrode de pH dans l'orifice de la base.
2.3 Installation de l'adaptateur d'alimentation
Insérer la pince d'adaptation adéquate dans la fente de l'adaptateur d'alimentation. L'alimentation
des balances Ohaus est aussi de 12 V, elle peut donc être utilisée pour le pH-mètre.
STARTER 3100 FR-7
2.4 Connexion des électrodes de pH
Il existe 3 supports d'électrodes : support pH (BNC), support Température (Cinch) et support REF
(banane 2 mm).
Pour le ST310, llectrode doit être connectée aux supports BNC et Cinch. Pour l'électrode de pH
2 en 1 (par exemple, ST210), seule la connexion BNC est nécessaire.
Le support REF est destiné à l'électrode de rérence séparée (par exemple, STREF01)
2.5 Support de réglage de l'angle de vue de l'appareil de mesure
La conception unique du STARTER 3100 concerne le support de l'appareil de mesure qui peut être
utilisé pour régler l'angle d'affichage en cas de travail sur une table élevée.
2.6 Guide rapide attaché
Un autre concept unique du STARTER 3100 concerne le Guide rapide attaché ; celui-ci est fixé en
partie inférieure du boîtier de l'appareil de mesure et peut être extrait d'un seul doigt.
FR-8 STARTER 3100
3 Configuration
3.1 finition de l'unité de température
Notez s'il vous plaît:
Si une électrode de température est utilisé, compensation automatique de température (ATC) et la
température de l'échantillon sont affichées sur l'écran. Vous pouvez alors choisir d'ignorer la
configuration MTC (ci-dessous).
Si le compteur ne tecte pas une électrode de température ou un n'est pas utilisé, l'appareil
passe automatiquement à la compensation manuelle de la température de mode (MTC) et MTC
apparaît sur l'écran.
MTC peut être défini comme suit:
Allumez l'appareil en appuyant sur Quitter.
Maintenir appuyé le bouton Mode/Configuration jusqu'à ce que l'icône de configuration
apparaisse à l'écran et que l'unité de température en cours clignote.
Appuyer sur le bouton Vers le haut ou sur le bouton Vers le bas pour commuter entre °C et
°F.
Appuyer sur le bouton Lecture pour confirmer votre sélection.
Ensuite
Continuer avec le réglage de la température MTC ou
Appuyer sur le bouton Lecture pour confirmer votre sélection.
Appuyer sur le bouton Quitter pour revenir à l'écran de mesure.
3.2 finition de la compensation manuelle de la température
Après avoir confirmé l'unité de température, la MTC (compensation manuelle de température)
apparaît sur l'affichage. Appuyer alors sur le bouton Vers le haut ou sur le bouton Vers le bas
pour augmenter ou diminuer la valeur de la température de votre échantillon. Appuyez sur le
bouton Lecture pour confirmer le glage ou appuyer sur le bouton Quitter pour revenir à l'écran
de mesure.
Le réglage de la valeur de la température MTC par défaut est de 25 °C (77 °F).
Remarque : °C = 5/9 (°F -32)
3.3 lection d'un groupe de tampons préfinis
Après avoir confirmé la valeur de la température de compensation MTC, il passe à la lection du
groupe de tampons. Appuyer sur le bouton Vers le haut ou sur le bouton Vers le bas pour
lectionner un groupe de tampons parmi quatre groupes (3.1.1). Appuyez sur le bouton Lecture
pour confirmer le glage ou appuyer sur le bouton Quitter pour quitter et revenir à l'écran de
mesure.
Le groupe de tampons par défaut est b1.
b1
1,68
4,01
7,00
10,01
STARTER 3100 FR-9
4 Fonctionnement du STARTER 3100
La procédure de mesure du pH est la suivante : les points clés sont les électrodes de pH, le
rinçage, l'étalonnage et le stockage. Après avoir introduit l'électrode dans la solution, il est
cessaire de l'homogéiser d'abord et ensuite de procéder à son utilisation.
a) Préparation de l'électrode de pH
b) préparation du tampon et étalonnage de l'électrode de pH
c) préparation de l'échantillon
d) Mesure du pH
e) Enregistrement ou impression du résultat des mesures
f) Rinçage de l'électrode de pH et stockage
Préparation de l'électrode de pH : l'électrode de pH doit être rincée avec de l'eau pure avant et
après l'utilisation. Vérifier si l'électrode est endommagée. (Prendre garde à l'ampoule en verre)
L'électrode de pH doit être rangée dans une bouteille de stockage ; la solution dans la bouteille
est une solution de KCl 3M. Après avoir plongé llectrode de pH dans l'échantillon ou dans la
solution tampon, l'utilisateur doit homogéiser pendant plusieurs secondes et attendre ensuite
30 à 60 secondes que le signal soit stable, et appuyer ensuite sur le bouton de fonctionnement
talonnage ou mesure).
AVERTISSEMENT Ne pas utiliser l'équipement dans des environnements
dangereux. L'équipement n'est pas proté contre les explosions.
AVERTISSEMENT Lors de l'utilisation de solvants et de produits chimiques, se
conformer aux instructions du fabricant des produits chimiques et aux règles
rales de sécurité d'un laboratoire.
4.1 Étalonnage
4.1.1 Groupe de tampons
Le STARTER 3100 peut effectuer des étalonnages à 1, 2 ou 3 points.
Il existe 3 groupes de tampons dans l'appareil de mesure, il est possible de sélectionner le groupe
de tampons prés (voir 4.3), le groupe de tampons par faut est b1 (normalisé États-Unis),
et la valeur du tampon sera automatiquement reconnue pendant l'étalonnage. Les 3 groupes de
tampons préfinis sont à 25 °C) :
Le STARTER 3100 corrige automatiquement la pendance à la température des valeurs de pH
du tampon qui sont données sur le tableau suivant : groupe de tampons b1.
Ceci signifie que si la température des solutions de tampons est de 15 °C, la valeur d'étalonnage
obtenue doit correspondre à pH 1,67 (tampon 1,68), pH 4,00 (tampon 4,01), pH 7,04 (tampon
7,00) et à pH 10,12 (tampon 10,01).
b1
1,68
4,01
7,00
10,01
b2
2,00
4,01
7,00
9.21
11,00
b3
1,68
4,00
6,86
9,18
12,46
FR-10 STARTER 3100
Étalonnage: l'électrode de pH doit être étalonnée avec une solution tampon à pH standard avant
d'effectuer une mesure correcte du pH. L'étalonnage doit afficher la valeur correcte du pH
lorsque l'appareil de mesure reçoit un signal en mV provenant de l'électrode de pH.
Pente : le coefficient linéaire entre mV et pH conformément à la valeur théorique (par
exemple -59,16mV/pH à 25 °C ce qui signifie une pente à 100 %),
calage : la valeur en mV lorsque la valeur du pH est égale à 7,00. (La valeur
théorique est de 0 mV),
4.1.2 alisation d'un étalonnage à 1 point
Lors de l'étalonnage, Ohaus recommande d'utiliser le mode automatique de point final. Après
avoir allumé l'appareil en marche, s'assurer que le haut de l'écran indique à s'assurer que le
compteur est en mode point automatique End.
Mode du sultat final : Deux modes de résultat final sont fournis : sultat final automatique et
sultat final manuel. En maintenant appuyé le bouton Lecture, il est possible de commuter d'un
mode à l'autre.
o Pour le sultat final manuel d'une mesure ou d'un étalonnage, il est cessaire
d'appuyer sur le bouton Lecture : la mesure de l'échantillon se fige, l'icône de
stabilité du sultat final clignote 3 fois et se fige à l'écran. L'appareil de
mesure ne peut pas atteindre automatiquement le résultat final.
o En mode de sultat final automatique, l'appareil de mesure cide si la mesure
est stable et assure alors automatiquement un sultat final, la mesure se fige et
clignote 3 fois et disparaît ; les icônes de sultat final clignotent 3 fois et se
figent à l'écran.
o Critère de stabilité : le signal de l'entrée du capteur peut ne pas être modifié de
plus de 0,1 mV en 6 secondes.
Remarque : Avec l'étalonnage à point unique, seul le calage est glé. Si le capteur a é
antérieurement étalonné par étalonnage à points multiples, la pente pdemment
stockée restera. Dans le cas contraire, la pente théorique de 100 % (-59,16 mV/pH)
sera utilisée.
5 °C
1,67
4.01
7,09
10,25
10 °C
1,67
4,00
7,06
10,18
15°C
1,67
4,00
7,04
10,12
20 °C
1,68
4,00
7,02
10,06
25 °C
1,68
4,01
7,00
10,01
30 °C
1,68
4,01
6,99
9,97
35 °C
1,69
4,02
6,98
9,93
40 °C
1,69
4,03
6,97
9,89
45 °C
1,70
4,05
6,97
9,86
50 °C
1,71
4,06
6,96
9,83
STARTER 3100 FR-11
Lorsque le STARTER 3100 est en mode de mesure du pH (voir 3.2), placer l'électrode de pH dans
un tampon d'étalonnage, homogéiser pendant 5 secondes, attendre 30 secondes, et
Appuyer sur le bouton Étalonnage, « Cal 1 » s'affiche en partie inférieure gauche de l'écran
et « Cal » clignote. et apparaissent en partie supérieure de l'écran, clignote pendant
l'étalonnage.
L'appareil de mesure atteint le sultat final conformément au mode du sultat final, la
valeur du pH du point dtalonnage (p. ex., 7,00) avec un affichage des températures à l'écran.
L'étalonnage à 1 point est terminé, 3 choix sont offerts pour la suite des choses :
Appuyer sur le bouton Étalonnage pour effectuer un étalonnage à 2 points.
Appuyer sur le bouton Lecture pour stocker l'étalonnage à 1 point et quitter, le calage et
la pente sont présentés sur l'affichage pendant 3 secondes pour ensuite revenir à lcran de
mesure.
Appuyer sur le bouton Quitter pour refuser l'étalonnage et revenir à l'écran de mesure.
4.1.3 alisation d'un étalonnage à 2 points
aliser un étalonnage à 1 point conformément à la description ci-dessus.
Rincer l'électrode de pH avec de l'eau pure et essuyer l'eau avec un mouchoir en papier.
Placer l'électrode dans le tampon d'étalonnage suivant, homogéiser et attendre, appuyer
ensuite sur le bouton Étalonnage, « Cal 2 » s'affiche en partie inférieure de l'écran et « Cal »
clignote. En partie supérieure de l'écran et apparaissent, clignote pendant
l'étalonnage.
L'appareil de mesure atteint un sultat final conformément au mode du sultat final, la
valeur du pH du point d'étalonnage (p. ex., 4,01) avec l'affichage de la température à l'écran.
L'étalonnage à 2 points est terminé, 3 choix sont dorénavant à disposition :
Appuyer sur le bouton Étalonnage pour réaliser un étalonnage à 3 points.
Appuyer sur le bouton Lecture pour stocker l'étalonnage à 2 points et quitter, le calage et
la pente sont présentés sur l'affichage pendant 3 secondes pour ensuite revenir à l'écran de
mesure.
Appuyer sur le bouton Quitter pour refuser l'étalonnage et revenir à l'écran de mesure.
Remarque : L'utilisation d'une électrode de température ou d'une électrode avec un capteur de
température intégré lectrode de pH 3 en 1) est recommandée. Si vous utilisez le mode MTC,
vous devez entrer la valeur correcte de température et conserver tous les tampons et solutions
d'échantillons à la température définie. (Voir la section 4.2)
4.2 Mesure d'un échantillon
4.2.1 Mesure du pH
Placer l'électrode de pH dans l'échantillon. Homogéiser pendant 5 secondes ; attendre 30
secondes.
Appuyer sur le bouton Lecture pour commencer la mesure du pH, s’affiche à l'écran.
clignote pendant la mesure.
Lorsque l'appareil de mesure atteint le sultat final, la valeur du pH avec la température
s'affichent à lcran.
FR-12 STARTER 3100
4.2.2 Mesure mV
Appuyer sur le bouton Mode pour commuter entre les modes de mesure du pH et de
mesure mV.
Appliquer la même procédure que pour la mesure du pH afin de aliser une mesure mV.
4.3 Mesure de température
Pour une précision supérieure, il est recommandé d'utiliser une électrode de température intégrée
ou séparée.
Si une électrode de température est utilisée, ATC et la température de l'échantillon sont
affichés.
Si l'appareil de mesure ne détecte pas d'électrode de température, il passe automatiquement
au mode manuel de compensation de température et MTC apparaît. La température MTC
doit être définie.
Remarque : Le STARTER 3100 accepte le capteur de température NTC 30 kΩ. Avec l'ATC
(compensation automatique de température) ou la MTC (compensation manuelle de température),
la compensation de température ne corrige que les modifications en sortie de l'électrode et non
pas celles de la solution en cours. Ceci signifie que l'appareil de mesure corrige le signal (mV)
provenant des électrodes de pH afin d'obtenir une plus grande précision de la valeur du pH
conformément à la température réelle.
4.4 Utilisation de la mémoire
4.4.1 Stockage d'une mesure
Le STARTER 3100 peut stocker jusqu'à 99 sultats finaux.
Appuyer sur le bouton Stockage lorsque la mesure atteint un sultat final. M01 indique
qu'un résultat a é stocké.
En appuyant sur le bouton Stockage lorsque M99 est affic, FUL (plein) apparaît pour indiquer
que la moire est pleine. Pour stocker d'autres données, vous devrez effacer la moire. (Voir
la section 3.4.3)
4.4.2 Rappel depuis la mémoire
Maintenir appuyé le bouton Rappel pour rappeler les valeurs stockées dans la moire
lorsque la mesure en cours atteint un résultat final.
Appuyer sur le bouton Vers le haut ou sur le bouton Vers le bas pour naviguer parmi les
sultats stockés. R01 à R99 indique quelsultat est en cours d'affichage.
Appuyer sur le bouton Quitter pour quitter.
4.4.3 Effacement de la mémoire
Appuyer sur le bouton Vers le haut ou sur le bouton Vers le bas pour naviguer parmi les
sultats stockés jusqu'à ce que «MRCL» apparaisse.
Appuyer sur le bouton Lecture, CLr clignote,
Deux choix se présentent alors :
Appuyer sur le bouton Lecture pour confirmer la suppression de toutes les données
stockées.
Appuyer sur le bouton Quitter pour revenir au mode de mesure sans effacer la mémoire.
STARTER 3100 FR-13
4.5 Impression
Si l'imprimante est connectée au STARTER 3100 (par exemple, SF-F40A, le moin vert signifie
la ussite de la connexion, le bit en baud de l'imprimante est de 9600 bps, 8 bits de données,
aucun bit de parité, 1 bit d'arrêt), une copie papier est automatiquement e après chaque
mesure de résultat final ou d'étalonnage.
Le format de la copie papier après une mesure de pH est le suivant :
sultat final, Valeur, Température, ATC/MTC
Auto EP, pH 4,01, 25,0 °C, MTC
Les détails de la deuxième ligne sont les suivants :
Auto EP, pH 4,01, 25,0 °C, MTC
| | | |----Compensation manuelle de température
| | |---------------- Valeur de température et unité C
| |------------------------- Valeur du pH
|---------------------------------------- Point de résultat final
Le format de la copie papier après une mesure mV est le suivant :
sultat final, Valeur, Température, ATC/MTC
EP manuel, 182 mV, 23,2 °C, ATC
Le format de sortie imprimée en cas de message d'erreur est :
sultat final, Valeur, Température, ATC/MTC
Erreur3
La copie papier d'un étalonnage à 2 points est :
Tampon1 : pH 4,01
mV1 : 178 mV
Temp.1 : 25,0 °C
Tampon2 : pH 7,00
mV2 : 3mV
Temp.2 : 25,0 °C
Pente : 99 %
calage : 5mV
ATC/MTC : MTC
La copie papier d'un étalonnage à 3 points est :
Tampon1 : pH 4,01
mV1 : 178 mV
Temp.1 : 25,0 °C
Tampon2 : pH 7,00
mV2 : 0mV
Temp.2 : 25,0 °C
Tampon3 : pH 9,21
mV3 : -130 mV
Temp.3 : 25,0 °C
Pente : 100 %
calage : 0mV
ATC/MTC : ATC
Impression depuis la moire En filant dans la moire, il est possible d'imprimer l'entrée en
cours d'examen en maintenant appuyé le bouton Stockage pendant 3 secondes. Le format de
sortie imprimée est le suivant :
Rappel 01 :
sultat final, Valeur, Température, ATC/MTC
EP manuel, pH 4,01, 35,6 °C, ATC
FR-14 STARTER 3100
5 Maintenance
5.1 Message d’erreur
Erreur 0
Erreur d'accèsmoire
initialisation aux valeurs usine
Erreur 1
Échec de l'auto-diagnostic
ter la procédure d'auto-diagnostic et arrêter
d'appuyer sur les cinq touches dans les deux
minutes.
Erreur 2
Valeurs mesurées hors plage
rifier si l'électrode est correctement connectée
et en place dans la solution échantillon.
Erreur 3
Température mesurée du
tampon hors plage (<5 ou
>40 °C)
Conserver la température du tampon de pH dans
la plage pour l'étalonnage
Erreur 4
calage hors plage
calage > 60m V ou < - 60
mV
S'assurer que le tampon de pH est correct et
cent. Nettoyer ou remplacer l'électrode de pH.
Erreur 5
Pente hors plage
S'assurer que le tampon est correct et récent.
Nettoyer ou remplacer l'électrode de pH.
Erreur 6
L'appareil de mesure ne peut
pas reconnaître le tampon
S'assurer que le tampon est correct et récent.
rifier si le tampon n'a pas é utilisé plusieurs
fois.
Erreur 9
L'ensemble des données
actuelles a dé é stocké
une fois
Une mesure de résultat final ne peut être stoce
qu'une seule fois. Réaliser une nouvelle mesure
pour le stockage.
Erreur
10
Température de l'échantillon
hors limites
rifier la température de l'échantillon, le capteur
de température.
En cas d'erreur, l'appareil de mesure émettra aussi un bip 3 fois à titre d'alerte.
Pour d'autres problèmes, contacter le représentant des ventes d'Ohaus.
5.2 Maintenance de l'appareil de mesure
Ne jamais dévisser les deux moitiés de l'enceinte !
Les instruments de la série STARTER 3100 ne cessitent aucune maintenance autre qu'un
essuyage occasionnel avec un chiffon humide et le remplacement des batteries vides.
L'enceinte est fabriquée en polystyrène-butadiène-acrylonitrile (ABS). Ce matériau est attaqué par
certains solvants organiques comme le toluène, le xyne et l'éthyl thyl tone (MEK). Tout
versement doit être imdiatement asséché.
5.3 Maintenance de l'électrode
S'assurer que l'électrode est remplie d'une solution électrolytique. Toujours stocker l'électrode
conformément au manuel d'instruction de l'électrode ; ne pas la laisser s'assécher.
STARTER 3100 FR-15
Si la réponse de l'électrode devient lente ou si la pente n'est pas suffisamment bonne, essayer de :
o Tremper l'électrode dans de HCl 0,1M pendant plus de 8 heures.
o Pour des contaminants huileux ou gras, graisser la membrane avec de la laine de
coton trempée dans de l'acétone ou dans une solution savonneuse.
Après avoir traité l'électrode, un nouvel étalonnage doit être réali. Si la pente de l'électrode n'est
toujours pas bonne, l'électrode doit être remplacée.
5.4 Auto-diagnostic
Lorsque l'appareil de mesure est en marche, appuyer et maintenir simultanément enfoncés
les boutons Lecture et Étalonnage jusqu'à ce que l'appareil de mesure affiche l'écran
complet. Chaque icône clignote l'une après l'autre.
De cette manière, il est possible de rifier si toutes les icônes sont correctement présentées.
L'étape suivante consiste à rifier que les touches fonctionnent correctement. Ceci cessite
l'interaction de l'utilisateur.
Lorsque b clignote, cinq icônes sont affichées.
Appuyer sur les cinq touches dans n'importe quel ordre. Chaque fois que vous appuyez sur
une touche, une icône disparaît de l'écran ; continuer d'appuyer sur les autres touches
jusqu'à ce que toutes les icônes aient disparu.
Lorsque l'auto-diagnostic est terminé avec succès, PAS apparaît. En cas d'échec de l'auto-
diagnostic, le message d'erreur Err 1 apparaît.
Remarque : Vous devez terminer d'appuyer sur les cinq touches dans les 2 minutes, sinon Err 1
apparaît et vous devrez réter la procédure.
5.5 cupération des paramètres usine
Lorsque l'appareil de mesure est à l'arrêt, maintenir simultanément appus les boutons
Lecture, Étalonnage et Quitter pendant 3 secondes, l'écran affiche et clignote,
signifie «initialisation ». Deux choix se présentent à l'utilisateur :
Appuyer sur le bouton Lecture pour une initialisation selon les paramètres usine
(MTC, pente et calage, etc.), affichage redémarrer ensuite l'appareil de
mesure.
Ou appuyer sur le bouton Quitter pour quitter le paramétrage, affichage et mettre
ensuite l'appareil de mesure à l'arrêt.
FR-16 STARTER 3100
6 DONNÉES TECHNIQUES
6.1 Spécifications
Conditions ambiantes
Utilisation en intérieur seulement
Altitude : Jusqu'à 2000 m
Plage de température spécifiée : 5 à 40 °C
Humidité : humidité relative maximum à 80 % avec des températures jusqu'à 30 °C diminuant
linéairement à 50 % d'humidité relative à 40 °C
Variations secteur jusqu'à ± 10 % de la tension nominale
Installation de catégorie II
Degré de pollution : 2
Fonctionnement assuré à température ambiante entre 5 et 40 °C
Modèle
ST3100
Plage de mesure
pH -2,00…16,00
1999…1999 mV
-5 °C…110 °C
solution
pH 0,01
1 mV
0,1 °C
Limites d'erreur
pH ± 0,01
± 1 mV
± 0,5 °C
Étalonnage
3 points
3 groupes de tampons préfinis
moire
99 mesures
Les dernières données d'étalonnage
Alimentation
Entrée Adaptateur CA : 100-240V ~ X,XA 50/60 Hz
Sortie Adaptateur CA : 12V X,XA
Dimensions/poids
220 (L) x 175 (P) x 78 (H)/ 0,75 kg
Affichage
ACL avec rétroéclairage
Entrée
BNC, impédance > 10e+12 Ω
Cinch, NTC 30 kΩ
Entrée de rence
Prise fiche banane 2 mm
Compensation de température
ATC et MTC
Enceinte
ABS
STARTER 3100 FR-17
7 Conformité
Ce produit est conforme à la directive EMC 2004/108/EC et à la directive de
basse tension 2006/95/EC. La Déclaration de conformité est disponible en ligne
sur europe.ohaus.com/europe/en/home/support/compliance.aspx.
Conformément à la directive européenne 2002/96/CE sur les déchets
d’équipements électriques et électroniques (WEEE), cet appareil ne doit pas
être mis au rebut comme des ordures ménagères. Ceci est également valable
pour les pays en dehors de l’UE, selon leurs impératifs spécifiques. Veuillez
mettre au rebut ce produit conformément à la réglementation locale au point de
collecte spécifié pour les équipements électriques et électroniques. En cas de
questions, veuillez contacter l’autorité responsable ou le distributeur auprès
duquel cet appareil a é acheté. Si cet appareil change de propriétaire (pour
des raisons personnelles ou professionnelles), cette consigne doit lui être
communiquée. Les instructions de mise au rebut en Europe sont disponibles en
ligne sur europe.ohaus.com/europe/en/home/support/weee.aspx. Nous vous
remercions de votre contribution à la protection de l’environnement.
Remarque FCC : Cet équipement a é testé et déclaré conforme aux limites des appareils
numériques de classe A, selon la Section 15 desgles du FCC (Commission férale des
communications). Ces limites ont pour objectif de fournir une protection raisonnable contre
des interférences dangereuses lorsque l'équipement est utilisé dans un environnement
commercial. Cet équipement gére, utilise et peut rayonner des hautes fréquences et s'il
n'est pas installé et utilisé conformément au guide de l'utilisateur, il peut gérer des
interférences préjudiciables aux communications radio. L'utilisation de cet équipement dans
une zone résidentielle risque de gérer des interférences préjudiciables, auquel cas
l'utilisateur se verra dans l'obligation de rectifier la situation à ses frais.
Enregistrement ISO 9001
En 1994, le Bureau Veritus Quality International (BVQI) a octroyé la certification
d'enregistrement ISO 9001 à Ohaus Corporation, États-Unis d'Amérique, confirmant que le
système de gestion de la qualité d'OHAUS était conforme aux spécifications normalisées de
l'ISO 9001. Le 21 juin 2012, Ohaus Corporation, États-Unis d'Amérique, a é-enregistrée
à la norme ISO 9001:2008.
FR-18 STARTER 3100
8 Groupe de tampons
Le STARTER 3100 corrige automatiquement la dépendance à la température des valeurs de pH du groupe de
tampons qui sont données sur les tableaux suivants : b2, b3, vous pouvez trouver b1 dans 3.1.1) .
Groupe de tampons b2 Normes européennes
Temp°C pH2,00 pH4,01 pH7,00 pH9,21 pH11,00
5 2,02 4,01 7,09 9,45 11,72
10 2,01 4,00 7,06 9,38 11,54
15 2,00 4,00 7,04 9,32 11,36
20 2,00 4,00 7,02 9,26 11,18
25 2,00 4,01 7,00 9,21 11,00
30 1,99 4,01 6,99 9,16 10,82
35 1,99 4,02 6,98 9,11 10,64
40 1,98 4,03 6,97 9,06 10,46
45 1,98 4,04 6,97 9,03 10,28
50 1,98 4,06 6,97 8,99 10,10
Groupe de tampons b3 JJG119
Temp °C pH1,68 pH4,00 pH6,86 pH9,18 pH12,46
5 1,67 4,00 6,95 9,39 13,21
10 1,67 4,00 6,92 9,33 13,01
15 1,67 4,00 6,90 9,28 12,82
20 1,68 4,00 6,88 9,23 12,64
25 1,68 4,00 6,86 9,18 12,46
30 1,68 4,01 6,85 9,14 12,29
35 1,69 4,02 6,84 9,11 12,13
40 1,69 4,03 6,84 9,07 11,98
45 1,70 4,04 6,83 9,04 11,83
50 1,71 4,06 6,83 9,02 11,70
STARTER 3100 FR-19
GARANTIE LIMITÉE
Ohaus garantit que ses produits sont exempts de fauts matériels et de fabrication à compter de
la date de livraison pendant toute la durée de la garantie. Pendant la riode de garantie, Ohaus
parera ou, à sa convenance, remplacera sans frais toutes les pièces déterminées défectueuses,
sous réserve que le produit soit retourné, fret payé d'avance, à Ohaus.
Cette garantie ne s'applique pas si le produit a subi des dommages suite à un accident ou un
susage, a é exposé à des matériaux radioactifs ou corrosifs, contient des matériaux
étrangers ayant tré à l'intérieur ou après une maintenance ou des modifications apportées
par quiconque autre qu’Ohaus. En l’absence d’une carte d'enregistrement de garantie dûment
remplie, la riode de garantie commence à la date de l'expédition au revendeur agréé. Aucune
autre garantie expresse ou implicite n'est offerte par Ohaus Corporation. En aucun cas, Ohaus
Corporation ne peut être tenu responsable des dommages indirects.
Dans la mesure où les lois régissant les garanties varient d'un État à un autre et d'un pays à un
autre, veuillez contacter Ohaus ou votre repsentant local agréé Ohaus pour de plus amples
informations.
Índice
1 Introdução ............................................................................ 1
1.1 Definição de sinais e símbolos de aviso ...................................... 1
1.2 Precauções de segurança .............................................................. 2
1.3 Display e controles ......................................................................... 3
2 Instalação ............................................................................. 5
2.1 Conteúdo da embalagem ............................................................... 5
2.2 Instalando o suporte de eletrodo independente .......................... 6
2.3 Instalando o adaptador de força .................................................... 6
2.4 Conecte os eletrodos de pH ........................................................... 7
2.5 Suporte do medidor para ajuste do ângulo de visão .................. 7
2.6 Guia rápido incluso ......................................................................... 7
3 Configuração ........................................................................ 8
3.1 Configurar a unidade de temperatura ........................................... 8
3.2 Configure a compensação manual de temperatura .................... 8
3.3 Selecionando um grupo predefinido de buffer ............................ 8
4 Funcionamento do STARTER 3100 .................................... 9
4.1 Calibração ........................................................................................ 9
4.1.1 Grupo de buffer ............................................................................. 9
4.1.2 Executando calibração de 1 ponto .............................................. 10
4.1.3 Executando calibração de 2 pontos ............................................. 11
4.2 Medição de amostra ....................................................................... 11
4.2.1 Medição de pH ............................................................................. 11
4.2.2 Medição de mV ............................................................................ 11
4.3 Medição de temperatura ................................................................ 11
4.4 Utilizando a memória .................................................................... 12
4.4.1 Armazenando uma leitura ........................................................... 12
4.4.2 Acessando a memória ................................................................ 12
4.4.3 Apagando a memória .................................................................. 12
4.5 Impressão ...................................................................................... 13
5 Manutenção ........................................................................ 14
5.1 Mensagens de erro........................................................................ 14
5.2 Manutenção do medidor ............................................................... 14
5.3 Manutenção do eletrodo ............................................................... 14
5.4 Autodiagnóstico ............................................................................ 15
5.5 Recuperando as configurações de fábrica ................................ 15
6 Especificações ................................................................... 16
7 Grupo de buffers ................................................................ 17
STARTER 3100 PT-1
1 Introdução
Grato por optar por um produto da OHAUS.
Favor ler completamente o manual antes de utilizar o medidor de pH de bancada STARTER 3100
para evitar funcionamento incorreto.
O STARTER 3100 possui excelente relação preço/performance, sendo projetado com muitos
recursos úteis, incluindo um suporte de eletrodo independente, um bepper para alerta de erro.
Também oferecemos outros acessórios, por exemplo, eletrodo de pH para diferentes aplicações
e soluções buffer.
Você achará muitas funcionalidades úteis nesses medidores. Algumas delas são:
LCD grande com iluminação com guia rápido incluso, de fácil operação pelo usuário
Ponto final automático/manual que pode auxiliar os usuários a congelar o valor de leitura
estável
Três grupos de buffer (US, EU e JJG) atendem a diferentes requisitos
O RS232 pode conectar uma impressora para impressão, 99 memórias de dados com os
dados da última calibração
1.1 Definição de sinais e símbolos de aviso
Notas de segurança o marcadas com palavras e mbolos de aviso. Estas mostram avisos e
questões de segurança. Ignorar as notas de segurança pode levar a lesões pessoais, danos ao
instrumento, avarias e resultados falsos.
Palavras de sinais
AVISO Para uma situação perigosa com risco dio, possivelmente
resultando em ferimentos ou morte se não for evitada.
CUIDADO Para uma situação perigosa com baixo risco, resultando em danos
ao aparelho ou a propriedade ou perda de dados, ou lesões, se
o for evitada.
Atenção Para uma informação importante a respeito do produto.
Nota Para obter informações úteis sobre o produto
mbolos de aviso
Risco geral
Risco de explosão
Risco de corrosão
Corrente alternativa
PT-2 STARTER 3100
Corrente direta
1.2 Precauções de segurança
AVISO: Leia todos os avisos de segurança antes de instalar, fazer conexões ou manutenção
deste equipamento. Não seguir estes avisos pode resultar em lesões pessoais ou danos à
propriedade. Reter todas as instruções para futura referência.
Verifique que a faixa de tensão de entrada impressa na etiqueta de dados e o tipo de plugue
de tomada estão de acordo com os padrões locais de adaptação.
Certifique-se de que o cabo de alimentação não represente um obsculo em potencial ou o
perigo de tropeçar.
Utilize o equipamento apenas em locais secos.
Seque qualquer líquido que derramar imediatamente. O instrumento não é à prova d'água.
Quando utilizar produtos químicos e solventes, siga as instruções do produtor químico e as
regras de segurança geral do laboratório.
Utilize somente acessórios aprovados.
Opere o equipamento somente em condições ambientais especificadas neste manual.
Desconecte o equipamento da fonte de alimentação para limpeza.
o opere o equipamento em ambientes perigosos ou instáveis.
Serviço deve ser realizado apenas por pessoal autorizado.
STARTER 3100 PT-3
1.3 Display e controles
Displays
1
Condição do eletrodo
Slope: mais de 95% e
offset: ± (0-15) mV
Condição do eletrodo é boa
Slope: menos que 90%
ou offset: ± (35-60) mV
Condição do eletrodo não é
boa e necessita limpeza
2
Ícone de estabilidade de ponto final ; ícone de ponto final automático
3
Ícone de medição - ; calibração ou medição em andamento
4
Ícone de calibração - ; calibração em andamento
5
Ícone de configuração - ; instrumento está no modo de configuração, pode configurar
temperatura (MTC), grupo de buffer, etc.
6
Leitura pH/mV ou slope em processo de calibração
7
Ponto de calibração / Grupo de Buffer / mero de memória / Índice de erro
8
Compensação automática de temperatura - ATC ; Compensação manual de temperatura -
MTC
9
Temperatura durante medição ou offset (mV) em processo de calibração
2
3
4
1
5
6
7
8
9
PT-4 STARTER 3100
Controles
Botão
Pressione e solte
Pressione e mantenha
pressionado por 3 segundos
- Iniciar ou terminar a medição
- Confirmar configuração,
armazenar valor inserido
- Ativar/desativar ponto final
automático
,
- Iniciar a calibração
- Acessar os últimos dados de
calibração: slope e offset
- Ativar o medidor
- Sair e retornar à tela de medição
- Desativar o medidor
- Armazenar leitura atual na
memória
- Aumentar valor durante ajuste
- Rolar para cima na memória
- Acessar dados armazenados
- Imprimir dados da memória
atual
- Alternar entre pH e mV
- Diminuir valor durante ajuste
- Rolar para baixo na memória
- Entrar no modo de
configuração
- Iniciar autodiagnóstico
Ativar/desativar a iluminação
do LCD
STARTER 3100 PT-5
2 Instalação
Com cuidado, abra a caixa.
2.1 Conteúdo da embalagem
O modelo ST3100-B (embalagem sica) possui os seguintes itens:
Além do conteúdo ST3100-B, a embalagem ST3100-F também incluiu o seguinte:
Além do conteúdo ST3100-B, a embalagem ST3100-H também incluiu o seguinte:
Cada buffer de pH deve ser dissolvido em 250ml de água pura ou água deionizada em um
frasco volumétrico. Você também pode pedir da Ohaus solução buffer em frasco.
Acessórios:
Modelo
Descrição
Item NO
ST310
Eletrodo de pH recarregável de plástico 3 em 1
83033965
ST210
Eletrodo de pH recarregável de plástico 2 em 1
83033966
ST320
Eletrodo de pH com gel plástico 3 em 1 (sem
necessidade de recarregar)
83033967
ST230
Eletrodo de pH de amostra lamacenta de vidro 2
em 1
83033968
STPURE
Eletrodo de pH de amostra de água pura de
vidro 2 em 1
83033969
STTEMP30
Eletrodo de Temperatura
83033970
STORP1
Eletrodo ORP de plástico gel
30038555
STORP2
Eletrodo ORP de vidro recarregável
30038553
STREF01
Eletrodo de referência Ag/AgCl
30059253
ST3100-B
Unidades
STARTER 3100
1
Porta eletrodo independente
1 unidade
Fonte de força 12V
1 unidade
Conjunto de Pó Buffer para pH (4.01, 7.00, 10.01)
1 unidade
Eletrodo de pH recarregável 3 em 1 ST310
1
Cobertura de proteção
1
Conjunto de Pó Buffer para pH (4.01, 7.00, 10.01)
1 unidade
Eletrodo de pH gel 3 em 1 ST320
1
Cobertura de proteção
1
PT-6 STARTER 3100
Conjunto de pó de buffer (4.01; 7.00; 10.0)
83033971
Buffer pH4.01 250ml
30065083
Buffer pH7.00 250ml
30065084
Buffer pH9.21 250ml
30065085
Buffer pH10.00 250ml
30065086
Eletrólito de Referência de eletrodo de pH
30059255
Solução de proteção de eletrodo de pH
30059256
Frasco de proteção de sensor de pH (10 na embalagem)
30064800
Suporte de eletrodo independente (3100/3100C)
30058733
Impressora SF40A
30045641
Cobertura de proteção
30058734
2.2 Instalando o suporte de eletrodo independente
Instale o braço do eletrodo na base,
Após a ajustar o botão de tensão em certa medida, você poderá levantar e abaixar
diretamente o braço. Quando colocamos um eletrodo de pH no braço e executamos
calibração ou medição, podemos colocar o frasco de armazenamento do eletrodo de pH no
orifício da base.
2.3 Instalando o adaptador de força
Insira o clipe adaptador correto na ranhura do adaptador de força. A fonte de força da balança
Ohaus também é 12V, podendo também ser utilizada para o medidor de pH.
Afrouxar
Apertar
STARTER 3100 PT-7
2.4 Conecte os eletrodos de pH
3 soquetes para o eletrodo. Soquete "pH" (BNC); Soquete TEMP. (Cinch) e Soquete REF
(banana 2 mm).
Para o eletrodo ST310, deve-se conectar o soquete BNC e Cinch. Para eletrodo pH 2 em 1
(ST210), só é necessário conectar o BNC.
Soquete REF para eletrodo de referência separado. (por exemplo, STREF01)
2.5 Suporte do medidor para ajuste do ângulo de visão
Um recurso exclusivo do STARTER 3100 é o suporte do medidor, que pode ser utilizado para
ajustar o ângulo de visão do display caso trabalhe em uma mesa alta.
2.6 Guia rápido incluso
Um outro recurso exclusivo do STARTER 3100 é o guia rápido incluso, que está afixado no
gabinete inferior do medidor, podendo utilizar o dedo para pressioná-lo para sair.
PT-8 STARTER 3100
3 Configuração
3.1 Configurar a unidade de temperatura
Por favor, note:
Se for utilizado um eletrodo de temperatura, compensão automática de temperatura (ATC) e a
temperatura da amostra o exibidos na tela. Em seguida, você pode optar por ignorar a
configuração MTC (abaixo).
Se o medidor o detecta um eletrodo de temperatura ou um o é usado, o medidor
automaticamente alterna para o modo de compensão de temperatura Manual (MTC) e MTC
aparece na tela.
MTC pode ser definido como segue:
Ligar o medidor pressionando Exit.
Pressionar e manter o botão Mode/Setup pressionado até o ícone de setup aparecer
no display e a unidade de temperatura atual piscar.
Pressionar o botão up ou down para intercalar entre °C e °F .
Pressionar o botão Read para confirmar sua seleção.
A seguir
Continue com a configuração da temperatura MTC ou
Pressione Read para confirmar a configuração
Pressione o botão Exit para retornar à tela de medição.
3.2 Configure a compensação manual de temperatura
Após confirmar a unidade de temperatura, aparecerá no display o valor de temperatura MTC
(Compensação Manual de Temperatura). A seguir, utilize o botão up ou down para aumentar ou
diminuir o valor até a temperatura de sua amostra. Pressione o botão Read para confirmar a
configuração ou pressione o botão Exit para retornar à tela de medição.
O valor padrão da temperatura MTC é 25 °C (77 °F).
Nota: °C = 5/9 (°F -32)
3.3 Selecionando um grupo predefinido de buffer
Após confirmar o valor da temperatura de compensação MTC, acessa-se a seleção de grupo de
buffer. Utilize botão up ou down para selecionar um grupo de buffer entre os quatro grupos
existentes (3.1.1). Pressione o botão Read para confirmar a configuração ou o boo Exit para
sair e retornar à tela de medição.
O grupo-padrão de buffer é b1.
b1
1.68
4.01
7.00
10.01
STARTER 3100 PT-9
4 Funcionamento do STARTER 3100
O procedimento de medição de pH é o seguinte; os pontos principais o o eletrodo de pH,
enxágue, calibração, armazenamento.
a) preparação do eletrodo de pH
b) preparação do buffer e calibração do eletrodo de pH
c) preparação da amostra
d) preparação do pH
e) registro ou impressão do resultado da medição
f) enxágue e armazenamento do eletrodo de pH
Preparação do eletrodo de pH: o eletrodo de pH deverá ser de enxaguado com água pura antes
e após a utilização. Verifique se o eletrodo está fisicamente danificado (tome cuidado com o
bulbo de vidro)
O eletrodo de pH deverá ser armazenado no frasco de armazenamento; a solução no frasco é
uma solução 3M KCl. Após o eletrodo de pH ser colocado na amostra ou solão buffer, o
usuário deverá agitar rios segundos e, a seguir, aguardar 30 a 60 segundos para que o sinal
se estabilize e, em seguida, pressionar o botão para funcionar (Calibração ou medição).
4.1 Calibração
4.1.1 Grupo de buffer
STARTER 3100 O STARTER 3100 pode executar calibrações de 1, 2 ou 3 pontos.
3 grupos de buffer no medidor; vo poderá selecionar o grupo de buffer que preferir (ver 4.3);
o buffer padrão é b1 US standards; o valor de buffer será automaticamente reconhecido durante
a calibração. Os 3 grupos de buffer predefinidos são (a 25 °C):
O STARTER 3100 automaticamente corrige a dependência de temperatura dos valores de pH de
buffer estabelecidos na seguinte tabela grupo de buffer b1.
b1
1.68
4.01
7.00
10.01
b2
2.00
4.01
7.00
9.21
11.00
b3
1.68
4.00
6.86
9.18
12.46
5 °C
1.67
4.01
7.09
10.25
10 °C
1.67
4.00
7.06
10.18
15 °C
1.67
4.00
7.04
10.12
20 °C
1.68
4.00
7.02
10.06
25 °C
1.68
4.01
7.00
10.01
30 °C
1.68
4.01
6.99
9.97
35 °C
1.69
4.02
6.98
9.93
40 °C
1.69
4.03
6.97
9.89
45 °C
1.70
4.05
6.97
9.86
50 °C
1.71
4.06
6.96
9.83
PT-10 STARTER 3100
Calibração: o eletrodo de pH precisa ser calibrado com solução buffer padrão de pH antes de
uma medição correta de pH. A calibração exibirá o valor correto de pH quando o medidor
receber o sinal do valor mV do eletrodo de pH.
Slope: coeficiente linear entre mV e pH de acordo com o valor teórico (por exemplo, -
59.16mV/pH @ 25 °C, que significa 100% slope);
Offset: o valor mV quando o valor de pH for 7.00. (Valor teórico é 0 mV);
4.1.2 Executando calibração de 1 ponto
Ao realizar a calibração, Ohaus recomenda o uso do modo de ponto de extremidade do Auto. De
pois de ligar otaxímetro, certifiquese o topo da tela mostra para garantir que o medidor está
no modo de ponto final de Auto.
Modo Endpoint: Temos dois modos endpoint - auto endpoint e manual endpoint. Ao
pressionar e manter pressionado o botão Read, você poderá alternar entre os 2 modos de ponto
final.
o Para executar o ponto final manualmente de uma medição ou calibração,
pressione o botão Read: a leitura da amostra congela; o ícone de estabilidade de
ponto final pisca 3 vezes e congela no display.
o Quando estiver no modo de ponto final automático, o medidor considerará se a
leitura es estável e, a seguir, executará automaticamente o ponto final; a leitura
congela e pisca 3 vezes e, em seguida, desaparece; os ícones de ponto final
automático piscam 3 vezes e congelam no display.
o Critério de estabilidade: o sinal de entrada do sensor o pode mudar em mais
de 0.1mV em 6 segundos.
Nota: Com a calibração de 1 ponto, somente o offset é ajustado. Se o sensor foi anteriormente
calibrado com calibração multiponto, o slope anteriormente armazenado permanecerá. Caso
contrário, será utilizado o slope 100 % teórico (-59.16 mV / pH).
Quando o STARTER 3100 estiver no modo de medição de pH (ver 3.2), coloque o eletrodo de
pH em um buffer de calibração e a seguir:
Pressione o botão Cal; Cal 1é exibido na parte inferior esquerda da tela e Cal pisca.
e aparecem na parte superior da tela, pisca durante a calibração.
O medidor alcança o ponto final de acordo com o modo endpoint, o valor de pH do ponto
de calibração (por exemplo, 7.00) com a exibição de temperatura na tela.
A calibração de 1 ponto terminou; agora, temos 3 escolhas a seguir:
Pressionar o botão Cal para executar a calibração de 2 pontos.
Pressionar o botão Read para armazenar a calibração de 1 ponto e sair; o offset e o slope
o mostrados no display por 3 segundos; a seguir, retorna para a tela de medição.
Pressionar o botão Exit para rejeitar a calibração e voltar à tela de medição.
STARTER 3100 PT-11
4.1.3 Executando calibração de 2 pontos
Execute a calibração de 1 ponto conforme descrita acima.
Enxágue o eletrodo de pH com água pura e enxugue com o papel toalha.
Coloque o eletrodo no próximo buffer de calibração e pressione o botão Cal, Cal 2 será
exibido no canto esquerdo inferior da tela e Cal piscará na parte superior da tela. e
aparecerão, piscará durante a calibração.
O medidor alcança o ponto final de acordo com o modo de ponto final, o valor de pH do
ponto de calibração (por exemplo, 4.01) com a exibição da temperatura na tela.
A calibração de 2 pontos está finalizada; agora temos 3 escolhas a seguir:
Pressionar o botão Cal para executar a calibração de 3 pontos.
Pressionar o botão Read para armazenar a calibração de 2 pontos e sair; o exibidos no
display o offset e o slope por 3 segundos e, a seguir, retorna à tela de medição.
Pressionar o botão Exit para rejeitar a calibração e retornar à tela de medição.
Nota: A utilização de um eletrodo de temperatura ou um eletrodo com o sensor de temperatura
integrado (eletrodo de pH 3 em 1) é recomendada. Se utilizar o modo MTC, deve inserir o valor
correto da temperatura e manter todas as soluções de amostra e buffer na temperatura ajustada.
(ver 4.2)
4.2 Medição de amostra
4.2.1 Medição de pH
Coloque o eletrodo de pH na amostra
Pressione o botão Read para iniciar a medição de pH; aparecerá no display. piscará
durante a medição.
Quando o medidor alcançar o ponto final, o valor de pH com a exibição de temperatura
serão mostrados na tela.
4.2.2 Medição de mV
Pressione o botão Mode para intercalar entre o modo de medição de pH e o modo de
medição de mV.
Siga o mesmo procedimento para medição de pH para executar a medição de mV.
4.3 Medição de temperatura
Para melhor precisão, recomendamos utilizar um eletrodo de temperatura integrado ou um
eletrodo de temperatura separado.
Se um eletrodo de temperatura for utilizado, ATC e a temperatura da amostra serão
exibidos.
Se o medidor não detectar um eletrodo temperatura, passará automaticamente para o modo
de compensação manual de temperatura e aparecerá MTC A temperatura MTC deverá ser
ajustada.
PT-12 STARTER 3100
Nota: O STARTER 3100 aceita sensor de temperatura NTC 30 . ATC (Compensação
Automática de Temperatura) ou MTC (Compensação Manual de Temperatura), a Compensação
de Temperatura somente corrige alteração na saída do eletrodo, o alteração na solução real.
Isso significa que o medidor corrige o sinal (mV) dos eletrodos de pH para garantir um valor de
pH mais preciso de acordo com a temperatura real.
4.4 Utilizando a memória
4.4.1 Armazenando uma leitura
O STARTER 3100 pode armazenar até 30 resultados de pontos finais.
Pressione o botão Store quando a medição alcançar o ponto final. M01 indica que um
resultado foi armazenado.
Se pressionar o botão Store quando M99 for exibido, FUL será exibido, indicando que a memória
está cheia. Para armazenar mais dados, você terá que liberar espaço na memória. (Ver 3.4.3)
4.4.2 Acessando a memória
Pressione e mantenha o botão Recall pressionado para acessar os valores armazenados na
memória quando a medição atual alcançar o ponto final.
Pressione o botão up ou down para rolar pelos resultados armazenados. R01 a R30
indicam que o resultado es sendo exibido.
Pressione o botão Exit para sair.
4.4.3 Apagando a memória
Pressione o botão up ou down para navegar pelos resultados armazenados até MRCL
aparecer.
Pressione o botão Read, CLr piscará;
Agora, temos duas escolhas:
Pressione o botão Read para confirmar o apagamento de todos os dados armazenados.
Pressione o botão Exit para retornar ao modo de medição sem apagar a memória.
STARTER 3100 PT-13
4.5 Impressão
Se a impressora estiver conectada ao STARTER 3100 (por exemplo, SF-F40A, a luz verde
significa sucesso da conexão), será gerada automaticamente uma impressão após cada
calibração ou medição com ponto final.
O formato da impressão após uma medição de pH será:
Ponto Final, Valor, Temperatura, ATC/MTC
Auto EP, 4.01pH, 25.0 °C, MTC
Os detalhes da segunda linha são:
Auto EP, 4.01pH, 25.0 °C , MTC
| | | |---- Compensão Manual de Temperatura
| | |---------------- Valor de temperatura e unidade °C
| |------------------------- Valor de pH
|---------------------------------------- Ponto Final Automático
O formato da impressão após uma medição de mV é:
Ponto Final, Valor, Temperatura, ATC/MTC
Manual EP, 182mV, 23.2 °C, ATC
A impressão em caso de erro de mensagem será:
Ponto Final, Valor, Temperatura, ATC/MTC
Error3
A impressão para uma calibração de 2 pontos será:
Buffer1: 4.01pH
mV1: 178mV
Temp.1: 25.0 °C
Buffer2: 7.00pH
mV2: 3mV
Temp.2: 25.0 °C
Slope: 99%
Offset: 5mV
ATC/MTC: MTC
A impressão para uma calibração de 3 pontos será:
Buffer1: 4.01pH
mV1: 178mV
Temp.1: 25.0 °C
Buffer2: 7.00pH
mV2: 0mV
Temp.2: 25.0 °C
Buffer3: 9.21pH
mV3: -130mV
Temp.3: 25.0 °C
Slope: 100%
Offset: 0mV
ATC/MTC: ATC
Quando navegar pela memória, você poderá imprimir a entrada atualmente exibida pressionando
e mantendo pressionado (3 segundos) o botão Store. O formato da impressão será o seguinte:
Recall 01:
Ponto Final, Valor, Temperatura, ATC/MTC
Manual EP, 4.01pH, 35.6 °C, ATC
PT-14 STARTER 3100
5 Manutenção
5.1 Mensagens de erro
Error 0
Erro de acesso à memória
Volta para as configurações de fábrica
Error 1
Falha de autodiagnóstico
Repita o procedimento de autodiagnóstico e
verifique se terminou de pressionar todas as
cinco teclas dentro de 2 minutos.
Error 2
Valores medidos fora de faixa
Verifique se o eletrodo está adequadamente
conectado e colocado na solução de amostra.
Error 3
Temperatura medida de
buffer fora da faixa (<5 ou
>40 °C)
Mantenha a temperatura de buffer de pH dentro
da faixa de calibração
Error 4
Offset fora da faixa
offset > 60mV ou < - 60 mV
Verifique se o buffer de pH está correto e é novo;
Limpe ou substitua o eletrodo de pH.
Error 5
Slope fora da faixa
Verifique se o buffer de pH está correto e é novo;
Limpe ou substitua o eletrodo de pH.
Error 6
Medidor não reconhece o
buffer
Verifique se o buffer é correto e é novo; verifique
se o buffer não foi utilizado mais do que uma vez.
Error 9
Conjunto de dados atuais já
foi armazenado antes
Uma leitura de ponto final somente pode ser
armazenada uma vez. Execute uma nova
medição para armazenar.
Error 10
Temperatura de amostra fora
da faixa
Verifique a temperatura de amostra, o sensor de
temperatura.
Se o erro ocorrer, o medidor também irá bipar três vezes para alerta.
Quaisquer outros problemas, contate o representante de vendas da OHAUS ou envie um email
5.2 Manutenção do medidor
Jamais desparafuse as duas metades do gabinete!
Os instrumentos da rie STARTER 3100 o requerem nenhuma manutenção, exceto limpeza
ocasional com tecido umedecido e a substituição das baterias esgotadas.
O gabinete é feito de acrilonitrilo butadieno estireno (ABS). Esse material é atacado por alguns
solventes orgânicos, por exemplo, tolueno, xileno e metiletilcetona (MEK). Qualquer respingo
deverá ser imediatamente limpo.
5.3 Manutenção do eletrodo
Verifique se o eletrodo está cheio de solução de eletrólito. Sempre armazene o eletrodo de
acordo com o seu manual de instrução e não deixe secar.
STARTER 3100 PT-15
Se a resposta do eletrodo tornar-se lenta ou se o slope não for suficientemente bom, tente:
o Deixar o eletrodo em imerso 0.1M HCl por mais de 8 horas.
o Para contaminante óleo ou gordura, desengordure a membrana com algodão saturado
em acetona ou solução saponácea.
Após o tratamento do eletrodo, deverá ser executada uma nova calibração. Se o slope do
eletrodo ainda não for bom, pode ser necessário substituir o eletrodo.
5.4 Autodiagnóstico
Quando o eletrodo estiver ativado, pressione e mantenha pressionado o botão Read e o
botão Cal simultaneamente até que o eletrodo exiba a tela cheia. Cada ícone piscará um
após o outro.
Dessa forma, você poderá verificar se todos os ícones o corretamente exibidos. A próxima
etapa é verificar se as teclas estão funcionando corretamente. Isso exige interação do usuário.
Quando b piscar, cinco ícones serão exibidos.
Pressione as cinco teclas em qualquer ordem. Cada vez que pressionar uma tecla, um ícone
desaparecerá da tela; continue pressionando as outras teclas até que todos os ícones
tenham desaparecido.
Quando terminar o autodiagnóstico com sucesso, aparecerá PAS. Se houver falha no
autodiagnóstico, a mensagem de erro Err 1 será exibida.
Nota: Você terá que terminar de pressionar todas as cinco teclas dentro de 2 minutos; caso
contrário, aparecerá a mensagem Err 1 e você terá que repetir o procedimento.
5.5 Recuperando as configurações de fábrica
Quando o medidor estiver desligado, pressione e mantenha pressionados os botões Read,
Cal e Exit em conjunto por 3 segundos; a tela exibirá e piscará, significando “Reset”.
A seguir, temos 2 escolhas:
Pressione o botão Read para voltar às configurações de brica (MTC, slope e offset,
etc.), exibirá ; a seguir, reinicie o medidor.
Ou pressione o botão Exit para sair da configuração; exibirá ; a seguir, desligue
o medidor.
PT-16 STARTER 3100
6 Especificações
STARTER 3100
Faixa de medição
-2.00…16.00 pH
–1999…1999 mV
-5 °C…110 °C
Resolução
0.01 pH
1 mV
0.1 °C
Limites de erro
± 0.01 pH
± 1 mV
± 0.5 °C
Calibração
3 pontos
3 grupos predefinidos de buffer
Memória
99 medições
Últimos dados de calibração
Alimentação
110-240V/50Hz, DC 12V
Tamanho/peso
Aproximadamente 220 L x 175 D x 78 A mm / 0.75 kg
Display
Cristal líquido com iluminação
Entrada
BNC, impedância > 10e+12 Ω
Cinch, NTC 30 kΩ
Entrada de referência
Soquete banana de 2 mm
Compensação de temperatura
ATC & MTC
Gabinete
ABS
STARTER 3100 PT-17
7 Grupo de buffers
O STARTER 3100 corrige automaticamente a dependência de temperatura do valor de pH do
grupo buffer estabelecido nas tabelas a seguir (b2, b3); você pode encontrar b1. (ver 3.1.1)
Grupo de buffer b2 Padrão Europa
Temp °C pH2.00 pH4.01 pH7.00 pH9.21 pH11.00
5 2.02 4.01 7.09 9.45 11.72
10 2.01 4.00 7.06 9.38 11.54
15 2.00 4.00 7.04 9.32 11.36
20 2.00 4.00 7.02 9.26 11.18
25 2.00 4.01 7.00 9.21 11.00
30 1.99 4.01 6.99 9.16 10.82
35 1.99 4.02 6.98 9.11 10.64
40 1.98 4.03 6.97 9.06 10.46
45 1.98 4.04 6.97 9.03 10.28
50 1.98 4.06 6.97 8.99 10.10
Grupo de buffer b3 JJG119
Temp °C pH1.68 pH4.00 pH6.86 pH9.18 pH12.46
5 1.67 4.00 6.95 9.39 13.21
10 1.67 4.00 6.92 9.33 13.01
15 1.67 4.00 6.90 9.28 12.82
20 1.68 4.00 6.88 9.23 12.64
25 1.68 4.00 6.86 9.18 12.46
30 1.68 4.01 6.85 9.14 12.29
35 1.69 4.02 6.84 9.11 12.13
40 1.69 4.03 6.84 9.07 11.98
45 1.70 4.04 6.83 9.04 11.83
50 1.71 4.06 6.83 9.02 11.70
STARTER 3100 Bench pH Meter EN-19
Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054 USA
Tel: +1 (973) 377-9000
Fax: +1 (973) 944-7177
With offices worldwide/Con oficinas en todo el mundo/Avec des bureaux dans
le monde entier/Com escritórios no mundo inteiro
www.ohaus.com
PN 30031621 E © Ohaus Corporation 2016, all rights reserved/todos los
derechos reservados/tous droits servés/todos os direitos reservados
*30031621*

Transcripción de documentos

STARTER 3100 Bench pH Meter Instruction Manual Manual de instrucciones del pHmetro de mesa STARTER 3100 STARTER 3100 PH-mètre de laboratoire Manuel d’instructions STARTER 3100 Medidor de pH de Bancada Manual de Instruções 1 TABLE OF CONTENTS INTRODUCTION ...................................................................1 1.1 1.2 1.3 2 INSTALLATION .....................................................................5 2.1 2.2 2.3 2.4 2.5 2.6 3 Package contents .............................................................................. 5 Installing the stand-alone electrode holder ....................................... 6 Installing the power adapter .............................................................. 6 Connect the pH electrodes ................................................................ 7 Meter stand for adjusting view angle................................................. 7 Attached quick guide ......................................................................... 7 SETUP ...................................................................................8 3.1 3.2 4 Definition of Signal Warnings and Symbols ...................................... 1 Safety Precautions ............................................................................ 2 Display and controls .......................................................................... 3 Set temperature unit and MTC value ................................................ 8 Selecting a predefined buffer group .................................................. 8 STARTER 3100 OPERATION ...............................................9 4.1 Calibration ......................................................................................... 9 4.1.1 Buffer group ................................................................................... 9 4.1.2 Performing 1-point calibration ..................................................... 10 4.1.3 Performing 2-point calibration ......................................................11 4.2 Sample measurement ..................................................................... 12 4.2.1 pH measurement ......................................................................... 12 4.2.2 mV measurement ........................................................................ 12 4.3 Temperature measurement ............................................................. 12 4.4 Using the memory ........................................................................... 12 4.4.1 Storing a reading ......................................................................... 12 4.4.2 Recalling from memory ............................................................... 12 4.4.3 Clearing the memory ................................................................... 13 4.5 Printing ............................................................................................ 13 5 MAINTENANCE ..................................................................15 5.1 5.2 5.3 5.4 5.5 6 TECHNICAL DATA ..............................................................17 6.1 6.2 7 Error message ................................................................................. 15 Meter maintenance.......................................................................... 15 Electrode maintenance ................................................................... 16 Self-diagnosis .................................................................................. 16 Recover factory settings .................................................................. 16 Specifications .................................................................................. 17 Compliance ..................................................................................... 18 BUFFER GROUP ................................................................19 STARTER 3100 Bench pH Meter 1 EN-1 INTRODUCTION Thank you for choosing OHAUS. Please read the manual completely before using the STARTER 3100 bench pH meter to ensure proper setup, operation and maintenance. STARTER 3100 has an excellent performance/price ratio and is designed with many useful features including a stand-alone electrode holder and a beeper for error alert. Other accessories such as pH electrodes and buffer solutions are also available. STARTER 3100 offers many practical features and functions such as:  Large backlit LCD with Quick Guide attached under the meter to assist with user operation  Stand-alone electrode holder for maximum flexibility  Electrode condition icon automatically showing you the health of your pH electrode  RS232 port and memory to store up to 99 measurements 1.1 Definition of Signal Warnings and Symbols Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results. Signal Words WARNING For a hazardous situation with medium risk, possibly resulting in injuries or death if not avoided. CAUTION For a hazardous situation with low risk, resulting in damage to the device or the property or in loss of data, or injuries if not avoided. Attention For important information about the product. Note For useful information about the product Warning Symbols General hazard Explosion hazard Corrosive hazard   Alternating current Direct current EN-2 1.2 STARTER 3100 Bench pH Meter Safety Precautions CAUTION: Read all safety warnings before installing, making connections, or servicing this equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain all instructions for future reference.           Verify that the input voltage range printed on the data label and the plug type matches the local AC power to be used. Make sure that the power cord does not pose a potential obstacle or tripping hazard. Use the equipment only in dry locations. Dry off any liquid spills immediately. The instrument is not watertight. When using chemicals and solvents, comply with the instructions of the chemical producer and the general lab safety rules. Use only approved accessories and peripherals. Operate the equipment only under ambient conditions specified in these instructions. Disconnect the equipment from the power supply when cleaning. Do not operate the equipment in hazardous or unstable environments. Service should only be performed by authorized personnel. STARTER 3100 Bench pH Meter 1.3 EN-3 Display and controls Displays 2 3 4 5 2 3 4 5 1 1 8 7 1 6 9 Electrode condition Slope: more than 95% and offset: ± (0-15) mV Electrode condition is good Slope: 90-95% or offset: ± (15-35) mV Electrode condition is acceptable Slope: less than 90% or offset: ± (35-60) mV Electrode condition is not good or needs cleaning 2 Endpoint stability icon 3 ; Auto endpoint icon Measurement icon - 4 Calibration icon - 5 Setup icon etc. 6 pH/mV reading or slope in calibration process 7 Calibration point 8 Auto temperature compensation - ATC ; Manual temperature compensation - MTC 9 Temperature during measurement or offset (mV) in calibration process ; means measurement or calibration is running when blinks ; means calibration in progress when display ; instrument is in the setup mode, can set temperature(MTC), buffer group / Buffer group /Memory number / Error index EN-4 STARTER 3100 Bench pH Meter Controls Button Press & release Press & hold for 3 seconds - Start or finish measurement - Confirm setting, store entered value - Turn auto endpoint on / off - Start calibration - Recall the latest calibration data : slope and offset - Meter turn on - Exit and return to measurement screen - Meter turn off - Store current reading to memory - Increase value during setting - Scroll up through the memory - Recall stored data - Print current memory data - Switch between pH and mV - Decrease value during setting - Scroll down through the memory - Enter setup mode , - Start self-diagnosis Turn on/turn off the backlight of the LCD STARTER 3100 Bench pH Meter 2 EN-5 INSTALLATION Carefully unpack the box. 2.1 Package contents The model ST3100-B (basic package) has the following items: ST3100-B Units STARTER 3100 1 Stand-alone electrode holder 1 set In use cover 1 12V Power supply 1 set In addition to ST3100-B content, the ST3100-F package also includes the following: pH Buffer Powder Set (4.01, 7.00, 10.01) 1 set ST310 3-in-1 refillable pH electrode 1 In addition to ST3100-B content, the ST3100-H package also includes the following: pH Buffer Powder Set (4.01, 7.00, 10.01) 1 set ST320 3-in-1 gel pH electrode 1 Each pH buffer powder should be dissolved in 250ml pure water or deionized water in a volumetric flask. Additional electrodes available include the following: Model Description ST310 3-in-1 plastic refillable pH Electrode ST210 2-in-1 plastic refillable pH Electrode ST320 3-in-1 plastic gel pH Electrode(no need to fill) ST230 2-in 1 glass muddy sample pH Electrode STORP1 Gel plastic ORP electrode STORP2 Refillable glass ORP electrode STTEMP30 Temperature Electrode P/N 83033965 83033966 83033967 83033968 30038555 30038553 83033970 EN-6 STARTER 3100 Bench pH Meter Buffers and Accessories: Buffer powder set (4.01; 7.00; 10.00) Buffer pH1.68 250ml Buffer pH4.01 250ml Buffer pH7.00 250ml Buffer pH10.01 250ml Buffer pH12.45 250ml Electrode holder stand alone (3100/3100C) In use cover 2.2 83033971 30100424 30100425 30100427 30100429 30100430 30058733 30058734 Installing the stand-alone electrode holder Install the electrode arm on the base, After adjusting the tension knob to some extent, you can move the upper arm up and down. When the pH electrode is installed on the arm, the storage bottle of pH electrode fits into the base hole. 2.3 Installing the power adapter Insert the right adapter clip into the power adapter slot. The Ohaus balance power supply is also 12V which could also be used for the pH meter. STARTER 3100 Bench pH Meter 2.4 EN-7 Connect the pH electrodes There are 3 sockets for electrode. “pH” socket (BNC); “TEMP.” Socket (Cinch) and “REF” Socket (2mm banana). For ST310 electrode connect to the BNC and Cinch socket. For 2-in-1 pH electrode (e.g. ST210) you only need to connect the BNC. REF socket is for separate reference electrode. (e.g.STREF1) 2.5 Meter stand for adjusting view angle One unique design of the STARTER 3100 is the meter stand, which can be used to adjust the view angle of the display in case of working on high table. 2.6 Attached quick guide Another unique design of the STARTER 3100 is the attached quick guide, the quick guide is attached into the bottom housing of the meter. EN-8 STARTER 3100 Bench pH Meter 3 SETUP 3.1 Set temperature unit and MTC value Please note: If a temperature electrode is used, Automatic Temperature Compensation (ATC) and the sample temperature are displayed on the screen. You may then choose to skip MTC setup (below). If the meter does not detect a temperature electrode or one is not used, the meter automatically switches to Manual Temperature Compensation (MTC) mode and MTC appears on the screen. MTC can be set as follows:  Power the meter on by pressing  Press and hold until the setup icon temperature unit blinks (°C or °F).  Press  Press Then or . appears on the display and the current to switch between °C and °F. to confirm your selection.  Continue with MTC temperature setting by using compensation accordingly  Press to confirm the setting  Press to return to the measurement screen. or to adjust temperature The default MTC temperature value setting is 25°C (77°F). Note: °C = 5/9 (°F - 32) 3.2 Selecting a predefined buffer group After confirming the MTC compensation temperature value, make the buffer group selection. Use or to select a buffer group among 3 buffer groups (4.1.1). Press confirm the setting or press The default buffer group is b1. b1 1.68 4.01 to leave to return to the measurement screen. 7.00 10.01 to STARTER 3100 Bench pH Meter 4 EN-9 STARTER 3100 OPERATION Standard procedure of pH measurement is as follows: a) pH electrode preparation b) buffer preparation and pH electrode calibration c) sample preparation d) pH measurement e) Record measurement results or print f) Rinse the pH electrode and properly store pH electrode preparation: pH electrode should be rinsed with pure water before and after using. Check if the electrode is physically damaged. (Be careful with the glass bulb) The pH electrode should be stored in the storage bottle; the solution in the bottle is 3M KCl solution. After placing the pH electrode into the sample or buffer solution, user should stir several seconds then wait 30 to 60 seconds for the signal to stabilize, and then press the button to operate (Calibration or measurement). WARNING Do not operate the equipment in hazardous environments. The equipment is not explosion protected. WARNING When using chemicals and solvents, comply with the instructions of the chemical producer and the general lab safety rules. 4.1 Calibration 4.1.1 Buffer group STARTER 3100 can perform 1-, 2- or 3- point calibrations. There are 3 buffer groups in the meter, you can select the buffer group you prefer (see 4.3), default buffer is b1 US standards; the buffer value will be automatically recognized during calibration. The 3 predefined buffer groups are (at 25°C): b1 1.68 4.01 7.00 10.01 b2 2.00 4.01 7.00 9.21 11.00 b3 1.68 4.00 6.86 9.18 12.46 STARTER 3100 automatically corrects for the temperature dependence of the buffer pH values given in the following table - buffer group b1. This means, if the buffer solutions’ temperature is 15°C, the calibration value you will get should be pH1.67 (buffer 1.68), pH4.00 (buffer 4.01), pH7.04(buffer 7.00) and pH10.12 (buffer 10.01). EN-10 STARTER 3100 Bench pH Meter 5 °C 1.67 4.01 7.09 10.25 10 °C 1.67 4.00 7.06 10.18 15 °C 1.67 4.00 7.04 10.12 20 °C 1.68 4.00 7.02 10.06 25 °C 1.68 4.01 7.00 10.01 30 °C 1.68 4.01 6.99 9.97 35 °C 1.69 4.02 6.98 9.93 40 °C 1.69 4.03 6.97 9.89 45 °C 1.70 4.05 6.97 9.86 50 °C 1.71 4.06 6.96 9.83 Calibration: pH electrodes need to be calibrated with pH standard buffer solution before a proper pH measurement can be made. Calibration is to display the right pH value when meter receive the mV value signal from the pH electrode. Slope: the linear coefficient between mV and pH according to theoretical value (e.g. 59.16mV/pH @ 25 °C which means 100% slope); Offset: the mV value when pH value is 7.00. (Theoretical value is 0 mV); 4.1.2 Performing 1-point calibration When performing calibration, Ohaus recommends using Auto End Point Mode. After powering the meter on, be sure the top of the screen shows to ensure the meter is in Auto End point Mode. Auto or Manual End point Mode: o o Press and hold to change the End Point Mode. When in Manual Mode, to manually reach a pH measurement or calibration value, o you need to press buttonwhen reading is stable and displays : then the sample reading or calibration value freezes, blinks 3 times and freezes on the display. When in Auto End Point Mode, the meter determines when the reading is stable then displays and locks the reading or calibration value automatically, the reading freezes and blinks 3 times then disappears; blinks 3 times and freezes on the display. Note: With the 1-point calibration only the offset is adjusted. If the sensor was previously calibrated with multi-point calibration the previously stored slope will remain. Otherwise theoretical 100% slope (-59.16 mV / pH) will be used. STARTER 3100 Bench pH Meter EN-11 When STARTER 3100 is in pH measurement mode (see 4.2); place the pH electrode in a calibration buffer, stir 5 seconds, wait for 30 seconds, then:   Press “Cal 1” displays on the bottom left of the screen and “Cal ” is blinking. and appear on the top of the screen, is blinking during calibration. The meter reaches endpoint automatically according to the preselected auto-endpoint mode, the calibration point pH value (e.g. 7.00) with the temperature display on the screen. The 1-point calibration is finished; There are now 3 options (OHAUS recommends conducting at least a 2 point calibration); 1. Press 2. Press to store the 1-point calibration and exit, the offset and the slope are shown on the display for 3 seconds then return to the measurement screen. 3. Press 4.1.3 to do the 2-point calibration. to reject the calibration, return to the measurement screen. Performing 2-point calibration  Perform 1-point calibration as described above.  Rinse the pH electrode with pure water and wipe off with tissue.  Place the electrode in the next calibration buffer, stir and wait, then press , “Cal 2” displays on the bottom left of the screen and “Cal ” is blinking. On the top of the screen and appear, is blinking during calibration. The meter reaches endpoint according to the endpoint mode, the calibration point pH value (e.g. 4.01) with the temperature display on the screen.  The 2-point calibration is finished. There are 3 options: 1. Press to do the 3-point calibration. 2. Press to store the 2-point calibration and exit, the offset and slope are shown on the display for 3 seconds then return to the measurement screen. 3. Press to reject the calibration, return to the measurement screen. Note: To perform a 3 point calibration, follow the instructions for performing a 2 point calibration using a third buffer. EN-12 STARTER 3100 Bench pH Meter 4.2 Sample measurement 4.2.1 pH measurement  Place the pH electrode in the sample. Stir for 5 seconds; wait for 30 seconds.  Press to start the pH measurement, appears on the display. is blinking during measurement. When meter reaches endpoint, the pH value with the temperature display on the screen.  4.2.2   mV measurement Press to switch between pH measurement mode and mV measurement mode. Follow the same procedure as for pH measurement to perform mV measurement. 4.3 Temperature measurement For better accuracy, we recommend to use either a built-in or a separate temperature electrode.  If a temperature electrode is used, ATC and the sample temperature are displayed.  If the meter does not detect a temperature electrode, it automatically switches to the manual temperature compensation mode and MTC appears. MTC temperature should be set. Note: STARTER 3100 accepts NTC 30 kΩ temperature sensor. ATC (Auto Temperature Compensation) or MTC (Manual Temperature Compensation), the Temperature Compensation only corrects for the change in the output of the electrode, not for the change in the actual solution. This means the meter corrects the signal(mV) from the electrode to get a more accurate pH value according to the real temperature. 4.4 Using the memory 4.4.1 Storing a reading The STARTER 3100 can store up to 99 endpoint results.  Press button has been stored. when the measurement reaches endpoint. M01 indicates that one result If you press when M99 is displayed, FUL displays to indicate the memory is full. To store further data you will have to clear the memory. (See 4.4.3) 4.4.2  Recalling from memory Press and hold to recall the stored values from memory when the current measurement reaches endpoint. STARTER 3100 Bench pH Meter  Press button or result is being displayed.  Press 4.4.3 EN-13 to scroll through the stored results. R01 to R99 indicates which to exit. Clearing the memory to scroll through the stored results until “MRCL” appears.  Pressing or  Press , CLr blinks; There are now 2 options:  Press to confirm the deletion of all the stored data.  Press to return to the measurement mode without deleting the memory. 4.5 Printing If the printer is connected to the STARTER 3100(e.g. SF-F40A, green light means connection success; printer baud rate 9600bps; 8 data bit; none parity bit; 1 stop bit), a print-out is automatically generated after each end pointed measurement or calibration. The format for the print-out following a pH measurement is: End Point, Value, Temp., ATC/MTC Auto EP, 4.01pH, 25.0 °C, MTC The details for the second line are: Auto EP, 4.01pH, 25.0 °C , MTC | | | |---- Manual Temperature Compensation | | |---------------- Temperature value and unit ºC | |------------------------- pH value |---------------------------------------- Auto End Point The format for the print-out following a mV measurement is: End Point, Value, Temp., ATC/MTC Manual EP, 182mV, 23.2 °C, ATC The print-out in case of an error message is: End Point, Value, Temp., ATC/MTC Error3 The print-out for a 2-point calibration is: Buffer1: 4.01pH mV1: 178mV Temp.1: 25.0 °C Buffer2: 7.00pH mV2: 3mV Temp.2: 25.0 °C EN-14 STARTER 3100 Bench pH Meter Slope: 99% Offset: 5mV ATC/MTC: MTC The print-out for a 3-point calibration is: Buffer1: 4.01pH mV1: 178mV Temp.1: 25.0 °C Buffer2: 7.00pH mV2: 0mV Temp.2: 25.0 °C Buffer3: 9.21pH mV3: -130mV Temp.3: 25.0 °C Slope: 100% Offset: 0mV ATC/MTC: ATC Printing from memory: When scrolling through the memory you can print the entry that is currently viewed by pressing and holding (3 seconds) the Recall 01: End Point, Value, Temp., ATC/MTC Manual EP, 4.01pH, 35.6 °C, ATC . The printout format is as follows: STARTER 3100 Bench pH Meter 5 MAINTENANCE 5.1 Error message EN-15 Error 0 Memory access error Reset to factory settings Error 1 Self-diagnosis failed Repeat the self-diagnosis procedure and make sure that you finish pressing all five keys within two minutes. Error 2 Measured values out of range Check if the electrode is properly connected and placed in the sample solution. Error 3 Measured buffer temperature out of range (<5 or >40 °C) Keep the pH buffer temperature within the range for calibration Error 4 Offset out of range offset > 60mV or < - 60 mV Make sure the pH buffer is correct and fresh; Clean or replace the pH electrode. Error 5 Slope out of range Make sure the buffer is correct and fresh; Clean or replace the pH electrode. Error 6 Meter cannot recognize the buffer Make sure the buffer is correct and fresh; check if the buffer has not been used more than once. Error 9 The current data set has already been stored once An endpoint reading can only be stored once. Perform a new measurement to store. Error 10 The sample temperature out of range Check the sample temperature, the temperature sensor. If an error happens, the meter will also beep 3 times to alert. For further technical support please contact Ohaus. (US please contact 1-800-672-7722). 5.2 Meter maintenance Never unscrew the two halves of the housing! The STARTER 3100 series instruments do not require any maintenance other than occasional wipe with a damp cloth and the replacement of depleted batteries. The housing is made of acrylonitrile butadiene styrene (ABS). This material is attacked by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage should be immediately wiped off. EN-16 5.3 STARTER 3100 Bench pH Meter Electrode maintenance Make sure the electrode is filled with electrolyte solution. Always store the electrode according to the electrode instruction manuals and do not allow it to dry out. If the electrode response becomes sluggish or the slope is not acceptable, try the following: o Soak the electrode in 0.1M HCl for at least 8 hours. o For fat or oil contaminant, degrease the membrane with cotton wool soaked in either acetone or a soap solution. After electrode treatment, a new calibration should be performed. If the electrode slope is still not acceptable, the electrode might need to be replaced. 5.4 Self-diagnosis  When the meter is on, press and hold and simultaneously until the meter displays the full screen. Each icon blinks one after the other. This way you may check whether all icons are correctly shown. The next step is to check that the keys are functioning correctly. This requires user interaction. When b blinks, five icons are displayed.  Press the five keys in any order. Each time you press a key an icon disappears from the screen, continue to press the other keys until all the icons have disappeared. When the self-diagnosis has been completed successfully, PAS appears. If self-diagnosis fails, error message Err 1 appears. Note: You have to finish pressing all five keys within 2 minutes, otherwise Err 1 appears and you will have to repeat the procedure. 5.5  Recover factory settings When the meter is off, press and hold seconds, the screen displays & & together for 3 and blinks, this means “Reset”. There are 2 options:  Press to reset factory settings (MTC, slope and offset, etc.), display then restart the meter.  Or press to quit the setting, display then turn off the meter. STARTER 3100 Bench pH Meter 6 TECHNICAL DATA 6.1 Specifications EN-17 Ambient conditions  Indoor use only  Altitude: Up to 2000 m  Specified Temperature range: 5 °C to 40 °C  Humidity: maximum relative humidity 80 % for temperatures up to 30 °C decreasing linearly to 50% relative humidity at 40 °C  Mains supply voltage fluctuations: up to ±10% of the nominal voltage  Installation category II  Pollution degree: 2  Operability is assured at ambient temperatures between 5 °C to 40 °C Model ST3100 Measuring range -2.00…16.00 pH –1999…1999 mV -5 °C…110 °C Resolution 0.01 pH 1 mV 0.1 °C Error limits ± 0.01 pH ± 1 mV ± 0.5 °C Calibration 3 points 3 predefined buffer groups Memory 99 measurements The last calibration data Power supply AC Adapter Input: 100-240V ~ X.XA 50/60 Hz AC Adapter Output: 12V  X.XA Size/weight 220 W x 175 D x 78 H mm / 0.75 kg Display Liquid crystal with backlight Input BNC, impedance > 10e+12 Ω Cinch, NTC 30 kΩ Reference input 2mm banana socket Temperature-compensation ATC & MTC Housing ABS EN-18 6.2 STARTER 3100 Bench pH Meter Compliance This product conforms to the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC. The Declaration of Conformity is available online at europe.ohaus.com/europe/en/home/support/compliance.aspx. In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. Disposal instructions in Europe are available online at europe.ohaus.com/europe/en/home/support/weee.aspx. Thank you for your contribution to environmental protection. FCC Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. ISO 9001 Registration In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the OHAUS quality management system is compliant with the ISO 9001 standards requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to the ISO 9001:2008 standard. STARTER 3100 Bench pH Meter 7 EN-19 BUFFER GROUP STARTER 3100 automatically correct for the temperature dependence of the buffer group pH value given in the following tables (b2, b3), you can find b1. (see 3.1.1) Buffer group b2 Europe standard Temp °C pH2.00 pH4.01 pH7.00 5 2.02 4.01 7.09 10 2.01 4.00 7.06 15 2.00 4.00 7.04 20 2.00 4.00 7.02 25 2.00 4.01 7.00 30 1.99 4.01 6.99 35 1.99 4.02 6.98 40 1.98 4.03 6.97 45 1.98 4.04 6.97 50 1.98 4.06 6.97 pH9.21 9.45 9.38 9.32 9.26 9.21 9.16 9.11 9.06 9.03 8.99 pH11.00 11.72 11.54 11.36 11.18 11.00 10.82 10.64 10.46 10.28 10.10 Buffer group b3 JJG119 Temp °C pH1.68 pH4.00 5 1.67 4.00 10 1.67 4.00 15 1.67 4.00 20 1.68 4.00 25 1.68 4.00 30 1.68 4.01 35 1.69 4.02 40 1.69 4.03 45 1.70 4.04 50 1.71 4.06 pH9.18 9.39 9.33 9.28 9.23 9.18 9.14 9.11 9.07 9.04 9.02 pH12.46 13.21 13.01 12.82 12.64 12.46 12.29 12.13 11.98 11.83 11.70 pH6.86 6.95 6.92 6.90 6.88 6.86 6.85 6.84 6.84 6.83 6.83 EN-20 STARTER 3100 Bench pH Meter LIMITED WARRANTY Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus. This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages. As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for further details. 1 Índice Introducción .........................................................................1 1.1 1.2 1.3 2 Instalación ............................................................................5 2.1 2.2 2.3 2.4 2.5 2.6 3 4.1.1 4.1.2 4.1.3 4.2 4.2.1 4.2.2 4.3 4.4 4.4.1 4.4.2 4.4.3 4.5 Grupo de tampones....................................................................... 9 Cómo realizar la calibración de un punto .................................... 10 Cómo realizar la calibración de dos puntos .................................11 Medición de la muestra ............................................................................ 11 Medición del pH............................................................................11 Medición de mV........................................................................... 12 Medición de la temperatura......................................................................12 Uso de la memoria ....................................................................................12 Almacenar una lectura ................................................................ 12 Recuperar desde la memoria ...................................................... 12 Borrar el contenido de la memoria .............................................. 12 Imprimir .....................................................................................................13 Mensajes de error .....................................................................................14 Mantenimiento del pHmetro .....................................................................14 Mantenimiento del electrodo ...................................................................14 Autodiagnóstico........................................................................................15 Recuperar los valores de fábrica .............................................................15 Datos técnicos ...................................................................16 6.1 7 8 Calibración ..................................................................................................9 Mantenimiento ...................................................................14 5.1 5.2 5.3 5.4 5.5 6 Ajustar la unidad de temperatura ..............................................................8 Ajustar la compensación de la temperatura manual ...............................8 Seleccionar un grupo predefinido de tampones ......................................8 Funcionamiento del STARTER 3100 ..................................9 4.1 5 Contenido del paquete ...............................................................................5 Instalación del soporte para electrodo autónomo ...................................6 Instalación del adaptador de corriente .....................................................6 Conexión de los electrodos pH .................................................................7 Soporte para el pHmetro para ajustar el ángulo de visión ......................7 Guía rápida incluida ....................................................................................8 Ajuste ....................................................................................8 3.1 3.2 3.3 4 Definición de palabras de advertencia y símbolos de advertencia ........1 Precauciones de seguridad .......................................................................2 Pantalla y controles ....................................................................................3 Especificaciones .......................................................................................16 Conformidad ......................................................................17 Grupo de tampones ...........................................................18 STARTER 3100 1 ES-1 Introducción Gracias por elegir un producto OHAUS. Para evitar el funcionamiento incorrecto, lea el manual completamente antes de utilizar el pHmetro de mesa STARTER 3100. Los STARTER 3100 tienen una excelente relación calidad/precio y están diseñados con muchas funciones útiles incluyendo un soporte para electrodo autónomo, una alarma para notificar un error. También están disponibles en OHAUS otros accesorios como varios electrodos pH para diferentes aplicaciones, y soluciones de tampón. En estos pHmetros encontrará funcionalidades muy útiles. Algunas de las muchas funciones son:     Gran pantalla LCD retroiluminada con guía rápida incluida para ayudar al usuario a trabajar con comodidad Punto final automático y manual que puede ayudar a los usuarios a fijar los valores de lectura estables 3 grupos de tampones predefinidos (EE.UU, UE y JJG) que cumplen con los diferentes requisitos RS232 puede conectar una impresora, memoria de 99 datos con los últimos datos de calibración 1.1 Definición de palabras de advertencia y símbolos de advertencia Las indicaciones de seguridad están marcadas con palabras de advertencia y símbolos de advertencia. Muestran cuestiones y advertencias de seguridad. Hacer caso omiso de las indicaciones de seguridad puede provocar lesiones personales, daños en el equipo, fallos de funcionamiento y resultados falsos. Palabras de advertencia ADVERTENCIA Para una situación de peligro con riesgo medio, que podría ocasionar lesiones o la muerte si no se evita. PRECAUCIÓN Para una situación de peligro con riesgo alto, que podría dañar el dispositivo o la propiedad o la pérdida de datos, o lesiones si no se evita. Atención Para información importante sobre el producto. Nota Para información útil sobre el producto ES-2 STARTER 3100 Símbolos de advertencia Peligro general Peligro de explosión Peligro de corrosión   1.2 Corriente alterna Corriente continua Precauciones de seguridad PRECAUCIÓN: Lea todas las advertencias de seguridad antes de la instalación, conexión o reparación del equipo. No cumplir con estas advertencias podría causar daños corporales y/o daños en la propiedad. Guarde las instrucciones para futuras consultas.           Compruebe que el rango de tensión de entrada impreso en la etiqueta de datos y el tipo de enchufe coincidan con la alimentación por CA utilizada. Asegúrese de que el cable de corriente no represente un posible obstáculo o peligro de tropiezo. Utilice el equipo solo en ambientes secos. Seque inmediatamente cualquier líquido derramado. El instrumento no es hermético. Cumpla las instrucciones del fabricante de productos químicos y las normas de seguridad generales de laboratorio cuando utilice productos químicos y disolventes. Utilice solo los accesorios y periféricos aprobados. Utilice el equipo solo en las condiciones ambientales especificadas en estas instrucciones. Desconecte el equipo de la fuente de alimentación cuando lo esté limpiando. No utilice el equipo en entornos peligrosos o inestables. El mantenimiento debe realizarse solamente por personal autorizado. STARTER 3100 1.3 ES-3 Pantalla y controles Pantallas 1 2 3 4 5 6 7 1 8 9 Condición del electrodo Pendiente: más del 95 % Pendiente: 90-95 % Pendiente: menos del 90 % y compensación: ± (0-15) mV o compensación: ± (15-35) mV o compensación: ± (35-60) mV La condición del electrodo es La condición del electrodo es La condición del electrodo no es buena aceptable buena y necesita limpieza 2 Icono de estabilidad del punto final 3 , icono de punto final automático Icono de medición ejecutando 4 Icono de calibración en curso 5 Icono de ajuste ; el instrumento está en modo ajuste, puede ajustar la temperatura (MTC), el grupo de tampones, etc. 6 Lectura de pH/mV o la pendiente en el proceso de calibración 7 Punto de calibración 8 Compensación automática de la temperatura - ATC, compensación manual de la temperatura - MTC 9 Temperatura durante la medición o la compensación (mV) en el proceso de calibración , cuando parpadea, indica que la medición o la calibración se está , cuando se muestra en la pantalla indica que la calibración está / grupo de tampones /Número de memoria / Índice de error ES-4 STARTER 3100 Controles Botón Pulsar y soltar Pulsar y mantener durante 3 segundos - Iniciar o finalizar la medición - Confirmar ajuste, almacenar un valor entero - Encender/apagar punto final automático - Iniciar calibración - Recuperar los últimos datos de calibración: pendiente, compensación - Encender el pHmetro Salir y volver a la pantalla de medición - Apagar el pHmetro - Almacenar en memoria lecturas actuales - Aumentar el valor durante la configuración - Desplazarse hacia arriba por la memoria - Cambiar entre pH y mV - Disminuir el valor durante la configuración - Desplazarse hacia abajo por la memoria Recuperar datos almacenados - Imprimir datos de memoria actuales , - Acceder al modo ajuste - Iniciar autodiagnóstico Encender/Apagar la retroiluminación de la LCD STARTER 3100 2 ES-5 Instalación Desempaquete cuidadosamente la caja. 2.1 Contenido del paquete El modelo ST3100-B (paquete básico) tiene los siguientes artículos: ST3100-B Unidades STARTER 3100 1 Soporte para electrodo autónomo 1 juego Cubierta en uso 1 Fuente de alimentación de 12 V 1 juego Además del contenido del ST3100-B, el paquete de ST3100-F también incluye lo siguiente: Juego de polvo para tampón de pH (4,01, 7,00, 10,01) 1 juego Electrodo de pH rellenable 3 en 1 ST310 1 Además del contenido del ST3100-B, el paquete de ST3100-H también incluye lo siguiente: Juego de polvo para tampón de pH (4,01, 7,00, 10,01) 1 juego Electrodo de pH gel 3 en 1 ST320 1 Se debe disolver calda tampón de pH para polvo en 250 ml de agua pura o desionizada en un matraz volumétrico. También puede solicitar el frasco con solución de tampón en Ohaus. Ahora hay disponibles más electrodos que incluyen lo siguiente: Modelo Descripción ST310 Electrodo de plástico para pH rellenable 3 en 1 ST210 Electrodo de plástico para pH rellenable 2 en 1 Electrodo de plástico para pH, con gel, 3 en 1 (no es ST320 necesario rellenarlo) ST230 Electrodo de vidrio para pH, muestras de lodo. 2 en 1 STORP1 Electrodo de plástico ORP, con gel STORP2 Electrodo de vidrio ORP, rellenable STTEMP30 Electrodo de temperatura P/N 83033965 83033966 83033967 83033968 30038555 30038553 83033970 ES-6 STARTER 3100 Tampones y accesorios: Juego de tampón para polvo (4,01; 7,00; 10,00) Tampón con pH 1,68, 250 ml Tampón con pH 4,01, 250 ml Tampón con pH 7,00, 250 ml Tampón con pH 10,01, 250 ml Tampón con pH 12,45, 250 ml Soporte para electrodo autónomo (3100/3100C) Cubierta en uso 2.2 83033971 30100424 30100425 30100427 30100429 30100430 30058733 30058734 Instalación del soporte para electrodo autónomo Coloque el brazo del electrodo a la base, Después de ajustar hasta cierto punto el mando de tensión, puede mover hacia arriba y hacia abajo la parte superior del brazo directamente. Al colocar el electrodo del pH en el brazo y realizar la calibración o medición, se puede colocar el frasco de almacenamiento del electrodo de pH en el hueco de la base. 2.3 Instalación del adaptador de corriente Inserte el clip adaptador correcto en la ranura del adaptador de corriente. La fuente de alimentación de la balanza Ohaus es también de 12 V que puede utilizarse también para el pHmetro. STARTER 3100 2.4 ES-7 Conexión de los electrodos pH Existen 3 enchufes por electrodo: el enchufe "pH" (BNC), “TEMP”, el enchufe (Cinch) y el enchufe “REF” (banana de 2 mm). Para el electrodo ST310 debe conectar el enchufe BNC y Cinch. Para el electrodo de pH 2 en 1 (p. ej. ST210) solo necesita conectar BNC. El enchufe REF es para el electrodo separado de referencia. (p. ej. STREF01) 2.5 Soporte para el pHmetro para ajustar el ángulo de visión Un diseño único del STARTER 3100 es el soporte para el pHmetro, que puede utilizarse para ajustar el ángulo de visión de la pantalla en caso de trabajar en mesas de laboratorio altas. ES-8 2.6 STARTER 3100 Guía rápida incluida Otro diseño único del STARTER 3100 es la guía rápida incluida. La guía rápida se incluye en el interior de la parte inferior de la carcasa del pHmetro, puede sacarla con los dedos. 3 Ajuste 3.1 Ajustar la unidad de temperatura Tenga en cuenta: Si se utiliza un electrodo de temperatura, compensación automática de temperatura (ATC) y la temperatura de la muestra se muestran en la pantalla. A continuación, puede optar por omitir la configuración MTC (abajo). Si el medidor no detecta un electrodo de temperatura o no se utiliza, el medidor cambia automáticamente al modo de compensación de temperatura manual (MTC) y MTC aparece en la pantalla. MTC se puede ajustar de la siguiente manera:  Poder del medidor pulsando Exit (Salir).  Mantenga pulsado el botón Mode/Setup (Modo/Ajuste) hasta que aparezca el icono en la pantalla y el estándar actual parpadee.  Pulse los botones Up (Arriba) o Down (Abajo) para cambiar entre °C y °F.  Pulse el botón Read (Leer) para confirmar su selección. Después  Continúe con la configuración de la temperatura MTC o  Pulse el botón Read (Leer) para confirmar su selección.  Pulse el botón Exit (Salir) volver a la pantalla de medición. 3.2 Ajustar la compensación de la temperatura manual Después de confirmar la unidad de la temperatura, se mostrará en la pantalla el valor de la temperatura MTC (Compensación de la temperatura manual). Después utilice los botones up (arriba) o down (abajo) para aumentar o disminuir el valor de la temperatura de su muestra. Pulse el botón Read (Leer) para confirmar el ajuste o pulse el botón Exit (Salir) para volver a la pantalla de medición. El ajuste del valor de la temperatura MTC por defecto es 25 °C (77 °F). Nota: °C = 5/9 (°F -32) 3.3 Seleccionar un grupo predefinido de tampones Después de confirmar el valor de la temperatura MTC de compensación, va hasta la selección del grupo de tampones. Utilice los botones Up (Arriba) o Down (Abajo) para seleccionar un grupo de tampones entre los 4 grupos de tampones (3.1.1). Pulse el botón Read (Leer) para confirmar el ajuste o pulse el botón Exit (Salir) para volver a la pantalla de medición. El grupo de tampones por defecto es b1. b1 1,68 4,01 7,00 10,01 STARTER 3100 4 ES-9 Funcionamiento del STARTER 3100 El procedimiento de la medición del pH es el siguiente, los puntos más importantes son el electrodo de pH, aclarado, calibración y almacenamiento. Después de colocar el electrodo en la solución, primero debemos mezclarlo y después realizar la operación. a) Preparar el electrodo de pH b) Preparar el tampón y la calibración del electrodo de pH c) Preparar la muestra d) Medir el pH e) Registrar o imprimir del resultado de la medición f) Enjuagar y almacenar el electrodo de pH Preparación del electrodo de pH: el electrodo de pH debe enjuagarse con agua pura antes y después de su uso. Compruebe si el electrodo está físicamente dañado (tenga cuidado con la pera de vidrio) El electrodo de pH debe almacenarse en el frasco de almacenamiento, la solución del frasco es una solución de KCl 3M. Después de colocar el electrodo de pH en la solución de la muestra o del tampón, el usuario debe mezclar durante varios segundos, después esperar de 30 a 60 segundos para que la señal se estabilice, y después pulsar el botón para comenzar a trabajar (calibración o medición). ADVERTENCIA No use el equipo en ambientes peligrosos. El equipo no está protegido contra explosiones. ADVERTENCIA Cumpla las instrucciones del fabricante de productos químicos y las normas de seguridad generales de laboratorio cuando utilice productos químicos y disolventes. 4.1 Calibración 4.1.1 Grupo de tampones El STARTER 3100 puede realizar calibraciones de 1-, 2- o 3- puntos. En el pHmetro hay 3 grupos de tampones, puede seleccionar el grupo de tampones que prefiera (consulte 4.3), el grupo de tampones por defecto es el estándar EE.UU b1 y durante la calibración se reconocerá automáticamente el valor del tampón. Los 3 grupos predefinidos de tampones son (a 25 °C): b1 1,68 4,01 7,00 10,01 b2 2,00 4,01 7,00 9,21 11,00 b3 1,68 4,00 6,86 9,18 12,46 El STARTER 3100 corrige automáticamente la dependencia de la temperatura de los valores de pH del tampón que figuran en la siguiente tabla, grupo del tampón b1. Esto quiere decir, que si la temperatura de la solución del tampón es de 15 °C, el valor de la calibración que obtendrá debería ser pH 1,67 (tampón 1,68), pH 4,00 (tampón 4,01), pH 7,04 (tampón 7,00) y pH 10,12 (tampón 10,01). ES-10 STARTER 3100 5 °C 1,67 4,01 7,09 10,25 10 °C 1,67 4,00 7,06 10,18 15 °C 1,67 4,00 7,04 10,12 20 °C 1,68 4,00 7,02 10,06 25 °C 1,68 4,01 7,00 10,01 30 °C 1,68 4,01 6,99 9,97 35 °C 1,69 4,02 6,98 9,93 40 °C 1,69 4,03 6,97 9,89 45 °C 1,70 4,05 6,97 9,86 50 °C 1,71 4,06 6,96 9,83 Calibración: Es necesario calibrar el electrodo de pH con la solución de tampón estándar de pH antes de realizar una medición de pH adecuada. La calibración sirve para mostrar el valor de pH correcto cuando el pHmetro recibe la señal del valor mV del electrodo de pH. Pendiente: el coeficiente lineal entre mV y pH según el valor teórico (p.ej. -59,16 mV/pH @ 25 °C lo que indica una pendiente del 100 %); Compensación: el valor de mV cuando el valor del pH es 7,00. (El valor teórico es 0 mV); 4.1.2 Cómo realizar la calibración de un punto Al realizar la calibración, Ohaus recomienda utilizar el modo de punto Colgar automáticamente. Después de encender el medidor, asegúrese de que la parte superior de la pantalla muestra para asegurar que el medidor está en modo de punto auto End. Modo de punto final: Hay dos modos de punto final, punto final automático y punto final manual. Al mantener pulsado el botón Read (Leer), puede cambiar entre los modos de punto final. o Para llevar al punto final manualmente una medición o calibración, tiene que pulsar el botón Read (Leer): La lectura de la muestra queda fija, el icono de estabilidad del punto final parpadeará 3 veces y se quedará fijo en la pantalla. El pHmetro no puede alcanzar automáticamente el punto final. o Cuando se encuentre en el modo de punto final automático, el pHmetro decide si la lectura es estable y después llega automáticamente al punto final, la lectura queda fija y parpadeará 3 veces y después desaparecerá, los iconos de punto final automático parpadearán 3 veces y quedarán fijos en la pantalla. o Criterio de estabilidad: La señal de la entrada del sensor puede no variar en más de 0,1 mV en 6 segundos. Nota: con la calibración de un punto solo se ajusta la compensación. Si el sensor se calibró previamente con calibración multipunto, se mantendrá la pendiente memorizada previamente. Si no, se utilizará la pendiente del 100% teórica (-59,16 mV / pH). STARTER 3100 ES-11 Cuando el STARTER 3100 se encuentre en modo medición del pH (consulte 3.2), coloque el electrodo de pH en un tampón de calibración, mezcle durante 5 segundos, espere 30 segundos y después:  Pulse el botón Cal. (Calibración), se mostrará “Cal 1” en la parte inferior izquierda de la pantalla y “Cal ” parpadeará. y aparecerán en la parte superior de la pantalla, parpadeará durante la calibración.  El pHmetro alcanzará el punto final según el modo de punto final, se mostrará en la pantalla el valor del pH del punto de calibración (p. ej. 7,00) junto con la temperatura. Ha finalizado la calibración de 1 punto, ahora existen 3 opciones para realizar a continuación:  Pulse el botón Cal (Calibración) para realizar la calibración de 2 puntos.  Pulse el botón Read (Leer) para almacenar la calibración de 1 punto y salir, durante 3 segundos se mostrarán en la pantalla la compensación y la pendiente y después volverá a la pantalla de medición.  Pulse el botón Exit (Salir) para rechazar la calibración, vuelva a la pantalla de medición. 4.1.3 Cómo realizar la calibración de dos puntos  Realizar la calibración de un punto como se describe anteriormente.  Enjuagar el electrodo de pH con agua pura y secar con un pañuelo de papel.  Colocar el electrodo en el siguiente tampón de la calibración, mezclar y esperar, después pulsar el botón Cal (Calibración), aparecerá “Cal 2” y “Cal ”parpadeará. En la parte superior de la pantalla aparecerá y , durante la calibración parpadeará .  El pHmetro alcanzará el punto final según el modo de punto final, se mostrará en la pantalla el valor del pH del punto de calibración (p. ej. 4,01) junto con la temperatura. Ha finalizado la calibración de dos puntos, ahora existen tres opciones para realizar a continuación:  Pulse el botón Cal (Calibración) para realizar la calibración de tres puntos.  Pulse el botón Read (Leer) para almacenar la calibración de dos puntos y salir, durante tres segundos se mostrarán en la pantalla la compensación y la pendiente y después volverá a la pantalla de medición.  Pulse el botón Exit (Salir) para rechazar la calibración, vuelva a la pantalla de medición. Nota: se recomienda el uso de un electrodo de temperatura o un electrodo con un sensor de temperatura integrado (electrodo de pH 3 en 1). Si utiliza el modo MTC, debe introducir el valor de temperatura correcto y mantener todas las soluciones de los tampones y de las muestras a la temperatura establecida. (consulte 4.2) 4.2 Medición de la muestra 4.2.1 Medición del pH  Coloque el electrodo de pH en la muestra.   Enjuague durante cinco segundos, espere 30 segundos. Pulse el botón Read (Leer) para comenzar la medición del pH. aparecerá en la pantalla . Durante la medición parpadeará. Cuando el pHmetro alcance el punto final, se mostrará en la pantalla el valor del pH junto con la temperatura. ES-12 4.2.2   STARTER 3100 Medición de mV Pulse el botón Mode (Modo) para cambiar entre el modo de medición del pH y el modo de medición de mV. Siga el mismo procedimiento de medición del pH para realizar una medición de mV. 4.3 Medición de la temperatura Para mayor precisión, recomendamos el uso de un electrodo de temperatura integrado o independiente.  Si se utiliza un electrodo de temperatura, se mostrarán en la pantalla la ATC y la temperatura de la muestra.  Si el pH metro no detecta un electrodo de temperatura, cambiará automáticamente al modo de compensación de temperatura manual y aparecerá MTC. Debe fijarse la temperatura MTC. Nota: el STARTER 3100 acepta el sensor de temperatura NTC 30 kΩ. ATC (compensación automática de la temperatura) o MTC (compensación manual de la temperatura), la compensación de la temperatura solo se corrige por el cambio de la salida del electrodo, no por el cambio en la solución actual. Esto quiere decir que el pHmetro corrige la señal (mV) de los electrodos pH para obtener una mayor exactitud del valor del pH según la temperatura real. 4.4 Uso de la memoria 4.4.1 Almacenar una lectura El STARTER 3100 puede almacenar hasta 99 resultados de punto final.  Pulse el botón Store (Almacenar) cuando la medición haya llegado al punto final. M01 indica que se ha almacenado un resultado. Si pulsa el botón Store (Almacenar) cuando se muestre M99, se mostrará FUL (Lleno) para indicar que no hay más espacio libre en la memoria. Para almacenar más datos deberá borrar el contenido de la memoria. (consulte 3.4.3). 4.4.2    Recuperar desde la memoria Cuando las mediciones lleguen al punto final, mantenga pulsado el botón Recall (Recuperar) para recuperar los valores de la memoria almacenados. Pulse los botones up (arriba) o down (abajo) para desplazarse por los resultados almacenados. De R01 a R99 indica qué resultado se está visualizando. Pulse el botón Exit (Salir) para salir. 4.4.3 Borrar el contenido de la memoria  Pulse los botones up (arriba) o down (abajo) para desplazarse por los resultados almacenados hasta que aparezca “MRCL” (Borrar memoria).  Pulse el botón Read (Leer), CLr (Borrar) parpadeará. Ahora tenemos dos opciones:  Pulsar el botón Read (Leer) para confirmar la eliminación de todos los datos almacenados.  Pulsar el botón Exit (Salir) para volver al modo de medición sin eliminar la memoria. STARTER 3100 4.5 ES-13 Imprimir Si una impresora está conectada al STARTER 3100 (p.ej. SF-F40A, la luz verde indica que la conexión se ha realizado con éxito, la tasa de baudios de la impresora es de 9600 bps, 8 bits de datos, sin bit de paridad, 1 bit de parada), se generará automáticamente una impresión después de que cada medición o calibración alcance el punto final. El formato para la impresión después de una medición de pH es: Punto final, valor, Temp., ATC/MTC EP automático, 4,01 pH, 25,0 °C, MTC Los detalles para la segunda línea son: EP automático, 4,01 pH, 25,0 °C, MTC | | | |---- compensación manual de la temperatura | | |----------------Valor de la temperatura y unidad °C | |------------------------- valor del pH |---------------------------------------- Punto final automático El formato para la impresión después de una medición de mV es: Punto final, valor, Temp., ATC/MTC EP manual, 182 mV, 23,2 °C , ATC La impresión en caso de un mensaje de error es: Punto final, valor, Temp., ATC/MTC Error3 La impresión para una calibración de 2 puntos con éxito es: Tampón1: 4,01 pH mV1: 178 mV Temp.1: 25,0 °C Tampón2: 7,00 pH mV2: 3 mV Temp.2: 25,0 °C Pendiente: 99 % Compensación: 5 mV ATC/MTC: MTC La impresión para una calibración de 3 puntos con éxito es: Tampón1: 4,01 pH mV1: 178 mV Temp.1: 25,0 °C Tampón2: 7,00 pH mV2: 0 mV Temp.2: 25,0 °C Tampón3: 9,21 pH mV3: -130 mV Temp.3: 25,0 °C Pendiente: 100% Compensación: 0 mV ATC/MTC: ATC Imprimir desde la memoria. Al desplazarse por la memoria puede imprimir la entrada que se está visualizando manteniendo pulsado el botón Store (Almacenar) durante 3 segundos. El formato de impresión es el siguiente: Recuperar 01: Punto final, valor, Temp., ATC/MTC EP automático, 4,01 pH, 35 °C, ATC ES-14 STARTER 3100 5 Mantenimiento 5.1 Mensajes de error Error 0 Error de acceso a la memoria Restablezca a los valores de fábrica Error 1 Fallo del autodiagnóstico Repita el procedimiento del autodiagnóstico y asegúrese de que finaliza pulsando las cinco teclas en el espacio de dos minutos. Error 2 Valores de medición fuera del intervalo Compruebe que el electrodo está correctamente conectado y colocado en la solución de muestra. Error 3 Temperatura de medición del tampón fuera del intervalo (<5 o >40 °C) Compensación fuera del intervalo Compensación > 60 mV o < 60 mV Pendiente fuera del intervalo Mantenga la temperatura del tampón del pH en el intervalo de calibración El pHmetro no reconoce el tampón Asegúrese de que el tampón es correcto y es reciente, compruebe que el tampón no se ha utilizado más de una vez. Una lectura de punto final solo puede almacenarse una vez. Realice una nueva medición para almacenar. Compruebe la temperatura de muestra, el sensor de temperatura Error 4 Error 5 Error 6 Error 9 Asegúrese de que el tampón pH es el correcto y es reciente. Limpie o reemplace el electrodo de pH. Asegúrese de que el tampón es el correcto y es reciente. Limpie o reemplace el electrodo de pH. El conjunto de datos actuales ya se ha almacenado una vez Error 10 La temperatura de muestra se encuentra fuera del intervalo Además, si ocurre un error, el pHmetro sonará tres veces para avisar. Para cualquier otro problema, póngase en contacto con el representante de ventas de Ohaus. 5.2 Mantenimiento del pHmetro ¡No desatornille nunca las dos mitades de la carcasa! Los equipos de la serie STARTER 3100 no necesitan ningún otro mantenimiento que una limpieza ocasional con un paño húmedo y el reemplazo de pilas gastadas. La carcasa está fabricada en ABS (acrilonitrilo-butadieno-estireno). Este material es sensible a algunos disolventes orgánicos como tolueno, xileno y metiletilcetona (MEK). Debe limpiarse inmediatamente cualquier derrame. 5.3 Mantenimiento del electrodo Asegúrese de que el electrodo esté lleno de solución electrolítica. Almacene siempre el electrodo según indique el manual de instrucciones del electrodo y no permita que se seque. STARTER 3100 ES-15 Si se ralentiza la respuesta del electrodo o la pendiente no es suficientemente buena, intente: o Sumergir el electrodo en 0,1 M HCI durante más de 8 horas. o Para los contaminantes de grasas y aceites, desengrase la membrana con un algodón empapado en acetona o una solución de jabón. Después del tratamiento del electrodo, debe realizarse una nueva calibración. Si la pendiente del electrodo sigue sin ser correcta, puede necesitar cambiar el electrodo. 5.4  Autodiagnóstico Cuando el pHmetro esté encendido, mantenga pulsado el botón Read (Leer) y el botón Cal (Calibración) simultáneamente hasta que el pHmetro muestre la pantalla completa. Los iconos parpadearán uno tras otro. De esta manera podrá comprobar si se muestran correctamente todos los iconos. El siguiente paso es el de comprobar que las teclas funcionan correctamente. Para esto es necesaria la interacción del usuario. Cuando b parpadee, se mostrarán cinco iconos.  Pulse las cinco teclas en cualquier orden. Cada vez que pulse una tecla, desaparecerá un icono de la pantalla, continúe pulsando las demás teclas hasta que hayan desaparecido todos los iconos. Cuando haya finalizado el autodiagnóstico con éxito, aparecerá PAS (Superado). Si el autodiagnóstico falla, aparecerá el mensaje de error Err 1 (Error 1). Nota: Debe finalizar pulsando las cinco teclas en el espacio de 2 minutos, de otra forma aparecerá el mensaje Err 1 (Error 1) y deberá repetir el procedimiento. 5.5  Recuperar los valores de fábrica Cuando el pHmetro esté apagado, mantenga pulsados a la vez durante 3 segundos los botones Read (Leer), Cal (Calibración) y Exit (Salir), la pantalla mostrará y parpadeará, lo que indicará que está "reiniciando". Tendremos entonces dos opciones:  Pulsar el botón Read (Leer) para restablecer los valores de fábrica (MTC, pendiente, compensación, etc.), se mostrará  después reinicie el pHmetro. O pulsar el botón Exit (Salir) para salir de la configuración, se mostrará apague el pHmetro. después ES-16 STARTER 3100 6 Datos técnicos 6.1 Especificaciones Condiciones ambientales  Solo para uso en interiores     Altitud: Hasta 2000 m Intervalo de temperatura especificado: 5 °C a 40 °C Humedad: humedad máxima relativa del 80 % para temperaturas de hasta 30 °C reducción de linealidad hasta una humedad relativa del 50 % a 40 °C  Fluctuaciones en el voltaje de la alimentación eléctrica: hasta ±10% del voltaje nominal Categoría de instalación II  Grado de polución: 2 El funcionamiento se asegura a temperaturas ambientales de entre 5 °C a 40 °C  Modelo ST3100 Intervalo de medición -2,00…16,00 pH –1999…1999 mV -5 °C…110 °C Resolución 0,01 pH 1 mV 0,1 °C Límites de error ± 0,01 pH ± 1 mV ± 0,5 °C Calibración 3 puntos 3 grupos predefinidos de tampones Memoria 99 mediciones El último dato de calibración Suministro de corriente Tamaño/peso Entrada para adaptador de CA: 100-240 V ~ X.XA 50/60 Hz Salida para adaptador de CA: 12 V  X.XA 220 (An) x 175 (Pr) x 78 (Al) (mm) / 0,75 Kg Pantalla Cristal líquido con retroiluminación Entrada BNC, impedancia > 10e+12 Ω Cinch, NTC 30 kΩ Entrada de referencia Enchufe tipo banana de 2 mm Compensación de temperatura ATC y MTC Carcasa ABS STARTER 3100 7 ES-17 Conformidad Este producto cumple con la directiva EMC 2004/108/CE y con la directiva de baja tensión 2006/95/CE. La declaración de conformidad está disponible a través de Internet en europe.ohaus.com/europe/en/home/support/weee.aspx. En cumplimiento con la Directiva europea 2002/96/CE sobre residuos de aparatos eléctricos y electrónicos (WEEE), este dispositivo no se puede eliminar como residuo doméstico. También se aplica en países fuera de la UE, según sus requisitos específicos. Elimine este producto según las normativas locales en el punto de recolección especificado para equipos eléctricos y electrónicos. Si tiene alguna pregunta, póngase en contacto con la autoridad responsable o el distribuidor del que adquirió este dispositivo. En caso de que este dispositivo se entregará a un tercero (para uso privado o profesional), se debe hacer referencia al contenido de esta regulación. Las instrucciones de eliminación para Europa están disponibles a través de Internet en europe.ohaus.com/europe/en/home/support/weee.aspx. Gracias por su contribución a la protección del medioambiente. Nota sobre la FCC: Este equipo ha sido sometido a pruebas y ha demostrado cumplir con los límites para dispositivos digitales de clase A, según la Parte 15 de las normas de la FCC. Estos límites están diseñados para proporcionar una protección razonable frente a interferencias perjudiciales cuando se utiliza el equipo en un entorno comercial. Este equipo genera, utiliza y puede emitir energía de radiofrecuencia y, si no se instala y utiliza según las instrucciones del manual puede provocar interferencias perjudiciales a las comunicaciones por radio. La utilización de este equipo en un área residencial puede provocar interferencias perjudiciales. En este caso, el usuario deberá corregir la interferencia por su cuenta. Registro en la ISO 9001 En 1994, OHAUS Corporation, EEUU, recibió el certificado de registro en la ISO 9001 del Bureau Veritus Quality International (BVQI), confirmando que el sistema de gestión de calidad de OHAUS cumple con los requisitos de la norma ISO 9001. El 21 de junio de 2012, OHAUS Corporation, EEUU, se registró nuevamente en la norma ISO 9001:2008. ES-18 8 STARTER 3100 Grupo de tampones El STARTER 3100 corrige automáticamente la dependencia de la temperatura del valor de pH del grupo del tampón que se facilita en las siguientes tablas (b2, b3), puede encontrar b1 (consulte 3.1.1) Grupo de tampón b2 Estándar europeo Temp °C pH 2,00 pH 4,01 pH 7,00 pH 9,21 5 2,02 4,01 7,09 9,45 10 2,01 4,00 7,06 9,38 15 2,00 4,00 7,04 9,32 20 2,00 4,00 7,02 9,26 25 2,00 4,01 7,00 9,21 30 1,99 4,01 6,99 9,16 35 1,99 4,02 6,98 9,11 40 1,98 4,03 6,97 9,06 45 1,98 4,04 6,97 9,03 50 1,98 4,06 6,97 8,99 pH 11,00 11,72 11,54 11,36 11,18 11,00 10,82 10,64 10,46 10,28 10,10 Grupo de tampón b3 JJG119 Temp °C pH 1,68 pH 4,00 pH 6,86 5 1,67 4,00 6,95 10 1,67 4,00 6,92 15 1,67 4,00 6,90 20 1,68 4,00 6,88 25 1,68 4,00 6,86 30 1,68 4,01 6,85 35 1,69 4,02 6,84 40 1,69 4,03 6,84 45 1,70 4,04 6,83 50 1,71 4,06 6,83 pH 12,46 13,21 13,01 12,82 12,64 12,46 12,29 12,13 11,98 11,83 11,70 pH 9,18 9,39 9,33 9,28 9,23 9,18 9,14 9,11 9,07 9,04 9,02 STARTER 3100 ES-19 GARANTÍA LIMITADA Los productos de Ohaus están en garantía por defectos en los materiales o de construcción desde la fecha de entrega hasta la duración del periodo de la garantía. Durante el periodo de validez de la garantía, Ohaus reparará o, según su criterio, sustituirá cualquier componente que sea defectuoso sin cargo adicional siempre que el producto se entregue, con flete pago, a Ohaus. Esta garantía no aplica si el producto ha recibido daños por accidentes o mal uso, ha sido expuesto a materiales radioactivos o corrosivos, ha recibido materiales extraños en su interior o como resultado del servicio o modificación por un tercero que no sea Ohaus. En lugar de una tarjeta de registro de garantía debidamente devuelta, el periodo de la garantía se iniciará en la fecha de envío al distribuidor autorizado. Ohaus Corporation no entrega otro tipo de garantía expresa o implícita. Ohaus Corporation no es responsable de daños indirectos. Puesto que la legislación sobre garantías difiere de un estado a otro y de un país a otro, póngase en contacto con Ohaus o con su distribuidor local de Ohaus para obtener más información. 1 Table des matières Introduction ..........................................................................1 1.1 1.2 1.3 2 Installation ............................................................................5 2.1 2.2 2.3 2.4 2.5 2.6 3 Contenu de l’emballage .................................................................. 5 Installation du porte-électrode autonome .................................... 6 Installation de l'adaptateur d'alimentation .................................... 6 Connexion des électrodes de pH .................................................. 7 Support de réglage de l'angle de vue de l'appareil de mesure .. 7 Guide rapide attaché....................................................................... 7 Configuration .......................................................................8 3.1 3.2 3.3 4 Définition des symboles et des indicateurs d'avertissements ... 1 Précautions de sécurité.................................................................. 2 Contrôles et affichage .................................................................... 3 Définition de l'unité de température .............................................. 8 Définition de la compensation manuelle de la température ....... 8 Sélection d'un groupe de tampons prédéfinis ............................. 8 Fonctionnement du STARTER 3100 ...................................9 4.1 Étalonnage ....................................................................................... 9 4.1.1 Groupe de tampons....................................................................... 9 4.1.2 Réalisation d'un étalonnage à 1 point ......................................... 10 4.1.3 Réalisation d'un étalonnage à 2 points ........................................11 4.2 Mesure d'un échantillon ............................................................... 11 4.2.1 Mesure du pH ...............................................................................11 4.2.2 Mesure mV .................................................................................. 12 4.3 Mesure de température ................................................................. 12 4.4 Utilisation de la mémoire .............................................................. 12 4.4.1 Stockage d'une mesure ............................................................... 12 4.4.2 Rappel depuis la mémoire .......................................................... 12 4.4.3 Effacement de la mémoire .......................................................... 12 4.5 Impression ..................................................................................... 13 5 Maintenance .......................................................................14 5.1 5.2 5.3 5.4 5.5 6 DONNÉES TECHNIQUES ..................................................16 6.1 7 8 Message d’erreur........................................................................... 14 Maintenance de l'appareil de mesure.......................................... 14 Maintenance de l'électrode .......................................................... 14 Auto-diagnostic ............................................................................. 15 Récupération des paramètres usine ........................................... 15 Spécifications ................................................................................ 16 Conformité..........................................................................17 Groupe de tampons ...........................................................18 STARTER 3100 1 FR-1 Introduction Merci d'avoir choisi un produit OHAUS. Lire le manuel complètement avant d'utiliser le pH-mètre de laboratoire STARTER 3100 afin d'éviter une exploitation inadéquate. Le STARTER 3100 possède un rapport performances/prix excellent et est conçu avec de nombreuses fonctions utiles, notamment un support d'électrode autonome et un bipeur pour les alertes d'erreurs. D'autres accessoires comme de nombreuses électrodes de pH pour différentes applications, et des solutions tampons en bouteilles sont aussi disponibles. Vous trouverez de nombreuses fonctionnalités utiles sur ces appareils de mesure. En voici quelques-unes ici présentées :  Grand ACL à rétroéclairage avec un Guide rapide attaché facilite l'exploitation par l'utilisateur  Résultat final automatique/manuel qui peut aider les utilisateurs à figer des mesures stables  3 groupes de tampons prédéfinis (États-Unis, EU et JJG) sont conformes à différentes spécifications  RS232 peut connecter une imprimante, 99 mémoires de données avec les dernières données d'étalonnage 1.1 Définition des symboles et des indicateurs d'avertissements Les remarques de sécurité sont marquées par des mots indicateurs et par des symboles d'avertissement. Elles présentent les problèmes et les avertissements de sécurité. Ne pas respecter les remarques de sécurité peut conduire à des accidents, à l'endommagement de l'instrument, à des dysfonctionnements et à des résultats erronés. Mots indicateurs AVERTISSEMENT Pour une situation dangereuse avec un risque moyen pouvant être à l'origine d'un accident ou d'un décès, s'il n'est pas évité. PRÉCAUTIONS Pour une situation dangereuse avec un faible risque pouvant être à l'origine de dommages au dispositif ou aux biens, à une perte des données, ou à un accident, s'il n'est pas évité. Attention Pour des informations importantes sur ce produit Remarque Pour des informations utiles sur le produit FR-2 STARTER 3100 Symboles d'avertissement Danger ! Risque d'explosion Danger, produits corrosifs   1.2 Tension alternative Courant continu Précautions de sécurité ATTENTION : Lire tous les avertissements de sécurité avant d’effectuer l’installation, les connexions ou la maintenance de cet équipement. Si ces avertissements ne sont pas respectés, des accidents et/ou des dommages matériels peuvent en résulter. Conserver toutes les instructions pour référence ultérieure.           Vérifier que la plage de tension d'entrée imprimée sur l'étiquette des données et le type de connecteur correspondent à l'alimentation CA locale à utiliser. S'assurer que le cordon d'alimentation n’est pas un obstacle potentiel ou ne présente aucun danger de chute. Utiliser l'équipement seulement dans des lieux secs. Sécher immédiatement tout déversement de liquide. L'instrument n'est pas étanche. Avec des produits chimiques et des solvants, se conformer aux instructions du fabricant des produits chimiques et aux règles générales de sécurité du laboratoire. Utiliser uniquement les accessoires et les périphériques approuvés. Faire fonctionner l'équipement uniquement dans les conditions ambiantes spécifiées dans ces instructions. Déconnecter l'équipement de l'alimentation électrique pour le nettoyer. Ne pas utiliser l'équipement dans des environnements dangereux ou instables. La maintenance doit être exclusivement exécutée par un personnel autorisé. STARTER 3100 1.3 FR-3 Contrôles et affichage Affichages 1 2 3 4 5 6 7 1 8 9 Condition de l'électrode Pente : plus de 95 % et décalage : ± (0-15) mV Bonne condition de l'électrode Pente : 90-95 % ou décalage : ± (15-35) mV Condition acceptable de l'électrode Pente : moins de 90 % ou décalage : ± (35-60) mV La condition de l'électrode n'est pas bonne ou nécessite un nettoyage 2 Icône de stabilité du résultat final 3 ; icône du résultat final automatique Icône de mesure - 4 Icône d'étalonnage - 5 Icône de configuration ; instrument en mode de configuration, possibilité de définir la température(MTC), le groupe de tampons, etc. 6 Lecture du pH/mV ou pente du processus d'étalonnage 7 Point d'étalonnage 8 Compensation automatique de température : ATC ; compensation manuelle de température : MTC 9 Température pendant la mesure ou décalage (mV) du processus d'étalonnage ; signifie mesure ou étalonnage en cours lorsqu'elle clignote ; signifie un étalonnage en cours lorsqu'elle clignote / Groupe tampon /Numéro mémoire / Index d'erreur FR-4 STARTER 3100 Contrôles Bouton Appuyer et relâcher Appuyer et maintenir enfoncé pendant 3 secondes - Démarrer ou terminer une mesure - Confirmation du réglage, stockage de la valeur entrée - Marche/arrêt du résultat final automatique - Démarrage de l'étalonnage - Rappel des dernières données d'étalonnage : pente et décalage - Appareil de mesure en marche - Quitter et revenir à l'écran de mesure - Appareil de mesure à l'arrêt - Stockage de la mesure en cours vers la mémoire - Augmentation de la valeur pendant les réglages - Défilement vers le haut de la mémoire - Commutation entre pH et mV - Diminution de la valeur pendant les réglages - Défilement vers le bas de la mémoire - Rappel des données stockées - Impression des données mémoire en cours , - Entrer dans le mode de configuration - Démarrage des auto-diagnostics Marche/arrêt du rétroéclairage de l'ACL STARTER 3100 2 FR-5 Installation Déballer soigneusement la boîte. 2.1 Contenu de l’emballage Le modèle ST3100-B (conditionnement de base) contient les articles suivants : ST3100-B Unités STARTER 3100 1 Porte-électrode autonome 1 ensemble Couvercle en utilisation 1 Alimentation 12 volts 1 ensemble En supplément au contenu du ST3100-B, le conditionnement du ST3100-F comprend aussi ce qui suit : Poudre pour tampon de pH (4,01, 7,00, 10,01) 1 ensemble Électrode de pH réutilisable ST310 3 en 1 1 En supplément au contenu du ST3100-B, le conditionnement du ST3100-H comprend aussi ce qui suit : Poudre pour tampon de pH (4,01, 7,00, 10,01) 1 ensemble Électrode de pH gel ST320 3 en 1 1 La poudre pour tampon de pH doit être dissoute dans 250 ml d'eau pure ou désionisée dans une fiole jaugée. Vous pouvez aussi commander auprès d'Ohaus des solutions de tampons en bouteille. Un plus grand nombre d'électrodes est désormais disponible et comprend ce qui suit : Modèle Description P/N ST310 Électrode de pH réutilisable en plastique 3 en 1 83033965 ST210 Électrode de pH à remplissage en plastique 2 en 1 83033966 Électrode de pH sous forme de gel en plastique 3 en 1 (nul ST320 besoin de remplissage) 83033967 ST230 Électrode de pH en verre pour échantillons troubles 2 en 1 83033968 STORP1 Électrode ORP en plastique pour gel 30038555 STORP2 Électrode réutilisable ORP en verre 30038553 STTEMP30 Électrode de température 83033970 FR-6 STARTER 3100 Pièces et accessoires : Poudre pour tampon de pH (4,01 ; 7,00 ; 10,00) Tampon pH 1,68 250 ml Tampon pH 4,01 250ml Tampon pH 7,00 250ml Tampon pH 10,01 250ml Tampon pH 12,45 250ml Porte-électrode autonome (3100/3100C) Couvercle en utilisation 2.2 83033971 30100424 30100425 30100427 30100429 30100430 30058733 30058734 Installation du porte-électrode autonome Installer le bras de l'électrode sur la base, Après avoir réglé le bouton de tension jusqu'à un certain niveau, il est possible de déplacer directement le bras supérieur vers le haut et vers le bas. En plaçant l'électrode de pH sur le bras et en réalisant un étalonnage ou une mesure, il est possible de ranger la bouteille de stockage de l'électrode de pH dans l'orifice de la base. 2.3 Installation de l'adaptateur d'alimentation Insérer la pince d'adaptation adéquate dans la fente de l'adaptateur d'alimentation. L'alimentation des balances Ohaus est aussi de 12 V, elle peut donc être utilisée pour le pH-mètre. STARTER 3100 2.4 FR-7 Connexion des électrodes de pH Il existe 3 supports d'électrodes : support pH (BNC), support Température (Cinch) et support REF (banane 2 mm). Pour le ST310, l'électrode doit être connectée aux supports BNC et Cinch. Pour l'électrode de pH 2 en 1 (par exemple, ST210), seule la connexion BNC est nécessaire. Le support REF est destiné à l'électrode de référence séparée (par exemple, STREF01) 2.5 Support de réglage de l'angle de vue de l'appareil de mesure La conception unique du STARTER 3100 concerne le support de l'appareil de mesure qui peut être utilisé pour régler l'angle d'affichage en cas de travail sur une table élevée. 2.6 Guide rapide attaché Un autre concept unique du STARTER 3100 concerne le Guide rapide attaché ; celui-ci est fixé en partie inférieure du boîtier de l'appareil de mesure et peut être extrait d'un seul doigt. FR-8 STARTER 3100 3 Configuration 3.1 Définition de l'unité de température Notez s'il vous plaît: Si une électrode de température est utilisé, compensation automatique de température (ATC) et la température de l'échantillon sont affichées sur l'écran. Vous pouvez alors choisir d'ignorer la configuration MTC (ci-dessous). Si le compteur ne détecte pas une électrode de température ou un n'est pas utilisé, l'appareil passe automatiquement à la compensation manuelle de la température de mode (MTC) et MTC apparaît sur l'écran. MTC peut être défini comme suit:  Allumez l'appareil en appuyant sur Quitter.  Maintenir appuyé le bouton Mode/Configuration jusqu'à ce que l'icône de configuration apparaisse à l'écran et que l'unité de température en cours clignote.  Appuyer sur le bouton Vers le haut ou sur le bouton Vers le bas pour commuter entre °C et °F.  Appuyer sur le bouton Lecture pour confirmer votre sélection. Ensuite  Continuer avec le réglage de la température MTC ou  Appuyer sur le bouton Lecture pour confirmer votre sélection.  Appuyer sur le bouton Quitter pour revenir à l'écran de mesure. 3.2 Définition de la compensation manuelle de la température Après avoir confirmé l'unité de température, la MTC (compensation manuelle de température) apparaît sur l'affichage. Appuyer alors sur le bouton Vers le haut ou sur le bouton Vers le bas pour augmenter ou diminuer la valeur de la température de votre échantillon. Appuyez sur le bouton Lecture pour confirmer le réglage ou appuyer sur le bouton Quitter pour revenir à l'écran de mesure. Le réglage de la valeur de la température MTC par défaut est de 25 °C (77 °F). Remarque : °C = 5/9 (°F -32) 3.3 Sélection d'un groupe de tampons prédéfinis Après avoir confirmé la valeur de la température de compensation MTC, il passe à la sélection du groupe de tampons. Appuyer sur le bouton Vers le haut ou sur le bouton Vers le bas pour sélectionner un groupe de tampons parmi quatre groupes (3.1.1). Appuyez sur le bouton Lecture pour confirmer le réglage ou appuyer sur le bouton Quitter pour quitter et revenir à l'écran de mesure. Le groupe de tampons par défaut est b1. b1 1,68 4,01 7,00 10,01 STARTER 3100 4 FR-9 Fonctionnement du STARTER 3100 La procédure de mesure du pH est la suivante : les points clés sont les électrodes de pH, le rinçage, l'étalonnage et le stockage. Après avoir introduit l'électrode dans la solution, il est nécessaire de l'homogénéiser d'abord et ensuite de procéder à son utilisation. a) Préparation de l'électrode de pH b) préparation du tampon et étalonnage de l'électrode de pH c) préparation de l'échantillon d) Mesure du pH e) Enregistrement ou impression du résultat des mesures f) Rinçage de l'électrode de pH et stockage Préparation de l'électrode de pH : l'électrode de pH doit être rincée avec de l'eau pure avant et après l'utilisation. Vérifier si l'électrode est endommagée. (Prendre garde à l'ampoule en verre) L'électrode de pH doit être rangée dans une bouteille de stockage ; la solution dans la bouteille est une solution de KCl 3M. Après avoir plongé l'électrode de pH dans l'échantillon ou dans la solution tampon, l'utilisateur doit homogénéiser pendant plusieurs secondes et attendre ensuite 30 à 60 secondes que le signal soit stable, et appuyer ensuite sur le bouton de fonctionnement (étalonnage ou mesure). AVERTISSEMENT Ne pas utiliser l'équipement dans des environnements dangereux. L'équipement n'est pas protégé contre les explosions. AVERTISSEMENT Lors de l'utilisation de solvants et de produits chimiques, se conformer aux instructions du fabricant des produits chimiques et aux règles générales de sécurité d'un laboratoire. 4.1 Étalonnage 4.1.1 Groupe de tampons Le STARTER 3100 peut effectuer des étalonnages à 1, 2 ou 3 points. Il existe 3 groupes de tampons dans l'appareil de mesure, il est possible de sélectionner le groupe de tampons préférés (voir 4.3), le groupe de tampons par défaut est b1 (normalisé États-Unis), et la valeur du tampon sera automatiquement reconnue pendant l'étalonnage. Les 3 groupes de tampons prédéfinis sont à 25 °C) : b1 1,68 4,01 7,00 10,01 b2 2,00 4,01 7,00 9.21 11,00 b3 1,68 4,00 6,86 9,18 12,46 Le STARTER 3100 corrige automatiquement la dépendance à la température des valeurs de pH du tampon qui sont données sur le tableau suivant : groupe de tampons b1. Ceci signifie que si la température des solutions de tampons est de 15 °C, la valeur d'étalonnage obtenue doit correspondre à pH 1,67 (tampon 1,68), pH 4,00 (tampon 4,01), pH 7,04 (tampon 7,00) et à pH 10,12 (tampon 10,01). FR-10 STARTER 3100 5 °C 1,67 4.01 7,09 10,25 10 °C 1,67 4,00 7,06 10,18 15°C 1,67 4,00 7,04 10,12 20 °C 1,68 4,00 7,02 10,06 25 °C 1,68 4,01 7,00 10,01 30 °C 1,68 4,01 6,99 9,97 35 °C 1,69 4,02 6,98 9,93 40 °C 1,69 4,03 6,97 9,89 45 °C 1,70 4,05 6,97 9,86 50 °C 1,71 4,06 6,96 9,83 Étalonnage: l'électrode de pH doit être étalonnée avec une solution tampon à pH standard avant d'effectuer une mesure correcte du pH. L'étalonnage doit afficher la valeur correcte du pH lorsque l'appareil de mesure reçoit un signal en mV provenant de l'électrode de pH. Pente : le coefficient linéaire entre mV et pH conformément à la valeur théorique (par exemple -59,16mV/pH à 25 °C ce qui signifie une pente à 100 %), Décalage : la valeur en mV lorsque la valeur du pH est égale à 7,00. (La valeur théorique est de 0 mV), 4.1.2 Réalisation d'un étalonnage à 1 point Lors de l'étalonnage, Ohaus recommande d'utiliser le mode automatique de point final. Après avoir allumé l'appareil en marche, s'assurer que le haut de l'écran indique à s'assurer que le compteur est en mode point automatique End. Mode du résultat final : Deux modes de résultat final sont fournis : résultat final automatique et résultat final manuel. En maintenant appuyé le bouton Lecture, il est possible de commuter d'un mode à l'autre. o Pour le résultat final manuel d'une mesure ou d'un étalonnage, il est nécessaire d'appuyer sur le bouton Lecture : la mesure de l'échantillon se fige, l'icône de stabilité du résultat final clignote 3 fois et se fige à l'écran. L'appareil de mesure ne peut pas atteindre automatiquement le résultat final. o En mode de résultat final automatique, l'appareil de mesure décide si la mesure est stable et assure alors automatiquement un résultat final, la mesure se fige et clignote 3 fois et disparaît ; les icônes de résultat final clignotent 3 fois et se figent à l'écran. o Critère de stabilité : le signal de l'entrée du capteur peut ne pas être modifié de plus de 0,1 mV en 6 secondes. Remarque : Avec l'étalonnage à point unique, seul le décalage est réglé. Si le capteur a été antérieurement étalonné par étalonnage à points multiples, la pente précédemment stockée restera. Dans le cas contraire, la pente théorique de 100 % (-59,16 mV/pH) sera utilisée. STARTER 3100 FR-11 Lorsque le STARTER 3100 est en mode de mesure du pH (voir 3.2), placer l'électrode de pH dans un tampon d'étalonnage, homogénéiser pendant 5 secondes, attendre 30 secondes, et  Appuyer sur le bouton Étalonnage, « Cal 1 » s'affiche en partie inférieure gauche de l'écran et « Cal » clignote. et apparaissent en partie supérieure de l'écran, clignote pendant l'étalonnage.  L'appareil de mesure atteint le résultat final conformément au mode du résultat final, la valeur du pH du point d'étalonnage (p. ex., 7,00) avec un affichage des températures à l'écran. L'étalonnage à 1 point est terminé, 3 choix sont offerts pour la suite des choses :  Appuyer sur le bouton Étalonnage pour effectuer un étalonnage à 2 points.  Appuyer sur le bouton Lecture pour stocker l'étalonnage à 1 point et quitter, le décalage et la pente sont présentés sur l'affichage pendant 3 secondes pour ensuite revenir à l'écran de mesure.  Appuyer sur le bouton Quitter pour refuser l'étalonnage et revenir à l'écran de mesure. 4.1.3 Réalisation d'un étalonnage à 2 points  Réaliser un étalonnage à 1 point conformément à la description ci-dessus.  Rincer l'électrode de pH avec de l'eau pure et essuyer l'eau avec un mouchoir en papier.  Placer l'électrode dans le tampon d'étalonnage suivant, homogénéiser et attendre, appuyer ensuite sur le bouton Étalonnage, « Cal 2 » s'affiche en partie inférieure de l'écran et « Cal » clignote. En partie supérieure de l'écran et apparaissent, clignote pendant l'étalonnage.  L'appareil de mesure atteint un résultat final conformément au mode du résultat final, la valeur du pH du point d'étalonnage (p. ex., 4,01) avec l'affichage de la température à l'écran. L'étalonnage à 2 points est terminé, 3 choix sont dorénavant à disposition :  Appuyer sur le bouton Étalonnage pour réaliser un étalonnage à 3 points.  Appuyer sur le bouton Lecture pour stocker l'étalonnage à 2 points et quitter, le décalage et la pente sont présentés sur l'affichage pendant 3 secondes pour ensuite revenir à l'écran de mesure.  Appuyer sur le bouton Quitter pour refuser l'étalonnage et revenir à l'écran de mesure. Remarque : L'utilisation d'une électrode de température ou d'une électrode avec un capteur de température intégré (électrode de pH 3 en 1) est recommandée. Si vous utilisez le mode MTC, vous devez entrer la valeur correcte de température et conserver tous les tampons et solutions d'échantillons à la température définie. (Voir la section 4.2) 4.2 Mesure d'un échantillon 4.2.1 Mesure du pH    Placer l'électrode de pH dans l'échantillon. Homogénéiser pendant 5 secondes ; attendre 30 secondes. Appuyer sur le bouton Lecture pour commencer la mesure du pH, s’affiche à l'écran. clignote pendant la mesure. Lorsque l'appareil de mesure atteint le résultat final, la valeur du pH avec la température s'affichent à l'écran. FR-12 4.2.2   STARTER 3100 Mesure mV Appuyer sur le bouton Mode pour commuter entre les modes de mesure du pH et de mesure mV. Appliquer la même procédure que pour la mesure du pH afin de réaliser une mesure mV. 4.3 Mesure de température Pour une précision supérieure, il est recommandé d'utiliser une électrode de température intégrée ou séparée.  Si une électrode de température est utilisée, ATC et la température de l'échantillon sont affichés.  Si l'appareil de mesure ne détecte pas d'électrode de température, il passe automatiquement au mode manuel de compensation de température et MTC apparaît. La température MTC doit être définie. Remarque : Le STARTER 3100 accepte le capteur de température NTC 30 kΩ. Avec l'ATC (compensation automatique de température) ou la MTC (compensation manuelle de température), la compensation de température ne corrige que les modifications en sortie de l'électrode et non pas celles de la solution en cours. Ceci signifie que l'appareil de mesure corrige le signal (mV) provenant des électrodes de pH afin d'obtenir une plus grande précision de la valeur du pH conformément à la température réelle. 4.4 Utilisation de la mémoire 4.4.1 Stockage d'une mesure Le STARTER 3100 peut stocker jusqu'à 99 résultats finaux.  Appuyer sur le bouton Stockage lorsque la mesure atteint un résultat final. M01 indique qu'un résultat a été stocké. En appuyant sur le bouton Stockage lorsque M99 est affiché, FUL (plein) apparaît pour indiquer que la mémoire est pleine. Pour stocker d'autres données, vous devrez effacer la mémoire. (Voir la section 3.4.3) 4.4.2    Maintenir appuyé le bouton Rappel pour rappeler les valeurs stockées dans la mémoire lorsque la mesure en cours atteint un résultat final. Appuyer sur le bouton Vers le haut ou sur le bouton Vers le bas pour naviguer parmi les résultats stockés. R01 à R99 indique quel résultat est en cours d'affichage. Appuyer sur le bouton Quitter pour quitter. 4.4.3  Rappel depuis la mémoire Effacement de la mémoire Appuyer sur le bouton Vers le haut ou sur le bouton Vers le bas pour naviguer parmi les résultats stockés jusqu'à ce que «MRCL» apparaisse.  Appuyer sur le bouton Lecture, CLr clignote, Deux choix se présentent alors :  Appuyer sur le bouton Lecture pour confirmer la suppression de toutes les données stockées.  Appuyer sur le bouton Quitter pour revenir au mode de mesure sans effacer la mémoire. STARTER 3100 4.5 FR-13 Impression Si l'imprimante est connectée au STARTER 3100 (par exemple, SF-F40A, le témoin vert signifie la réussite de la connexion, le débit en baud de l'imprimante est de 9600 bps, 8 bits de données, aucun bit de parité, 1 bit d'arrêt), une copie papier est automatiquement générée après chaque mesure de résultat final ou d'étalonnage. Le format de la copie papier après une mesure de pH est le suivant : Résultat final, Valeur, Température, ATC/MTC Auto EP, pH 4,01, 25,0 °C, MTC Les détails de la deuxième ligne sont les suivants : Auto EP, pH 4,01, 25,0 °C, MTC | | | |----Compensation manuelle de température | | |---------------- Valeur de température et unité C | |------------------------- Valeur du pH |---------------------------------------- Point de résultat final Le format de la copie papier après une mesure mV est le suivant : Résultat final, Valeur, Température, ATC/MTC EP manuel, 182 mV, 23,2 °C, ATC Le format de sortie imprimée en cas de message d'erreur est : Résultat final, Valeur, Température, ATC/MTC Erreur3 La copie papier d'un étalonnage à 2 points est : Tampon1 : pH 4,01 mV1 : 178 mV Temp.1 : 25,0 °C Tampon2 : pH 7,00 mV2 : 3mV Temp.2 : 25,0 °C Pente : 99 % Décalage : 5mV ATC/MTC : MTC La copie papier d'un étalonnage à 3 points est : Tampon1 : pH 4,01 mV1 : 178 mV Temp.1 : 25,0 °C Tampon2 : pH 7,00 mV2 : 0mV Temp.2 : 25,0 °C Tampon3 : pH 9,21 mV3 : -130 mV Temp.3 : 25,0 °C Pente : 100 % Décalage : 0mV ATC/MTC : ATC Impression depuis la mémoire En défilant dans la mémoire, il est possible d'imprimer l'entrée en cours d'examen en maintenant appuyé le bouton Stockage pendant 3 secondes. Le format de sortie imprimée est le suivant : Rappel 01 : Résultat final, Valeur, Température, ATC/MTC EP manuel, pH 4,01, 35,6 °C, ATC FR-14 STARTER 3100 5 Maintenance 5.1 Message d’erreur Erreur 0 Erreur d'accès mémoire Réinitialisation aux valeurs usine Erreur 1 Échec de l'auto-diagnostic Répéter la procédure d'auto-diagnostic et arrêter d'appuyer sur les cinq touches dans les deux minutes. Erreur 2 Valeurs mesurées hors plage Vérifier si l'électrode est correctement connectée et en place dans la solution échantillon. Erreur 3 Température mesurée du tampon hors plage (<5 ou >40 °C) Décalage hors plage décalage > 60m V ou < - 60 mV Pente hors plage Conserver la température du tampon de pH dans la plage pour l'étalonnage Erreur 6 L'appareil de mesure ne peut pas reconnaître le tampon Erreur 9 L'ensemble des données actuelles a déjà été stocké une fois Température de l'échantillon hors limites S'assurer que le tampon est correct et récent. Vérifier si le tampon n'a pas été utilisé plusieurs fois. Une mesure de résultat final ne peut être stockée qu'une seule fois. Réaliser une nouvelle mesure pour le stockage. Vérifier la température de l'échantillon, le capteur de température. Erreur 4 Erreur 5 Erreur 10 S'assurer que le tampon de pH est correct et récent. Nettoyer ou remplacer l'électrode de pH. S'assurer que le tampon est correct et récent. Nettoyer ou remplacer l'électrode de pH. En cas d'erreur, l'appareil de mesure émettra aussi un bip 3 fois à titre d'alerte. Pour d'autres problèmes, contacter le représentant des ventes d'Ohaus. 5.2 Maintenance de l'appareil de mesure Ne jamais dévisser les deux moitiés de l'enceinte ! Les instruments de la série STARTER 3100 ne nécessitent aucune maintenance autre qu'un essuyage occasionnel avec un chiffon humide et le remplacement des batteries vides. L'enceinte est fabriquée en polystyrène-butadiène-acrylonitrile (ABS). Ce matériau est attaqué par certains solvants organiques comme le toluène, le xylène et l'éthyl méthyl cétone (MEK). Tout déversement doit être immédiatement asséché. 5.3 Maintenance de l'électrode S'assurer que l'électrode est remplie d'une solution électrolytique. Toujours stocker l'électrode conformément au manuel d'instruction de l'électrode ; ne pas la laisser s'assécher. STARTER 3100 FR-15 Si la réponse de l'électrode devient lente ou si la pente n'est pas suffisamment bonne, essayer de : o Tremper l'électrode dans de HCl 0,1M pendant plus de 8 heures. o Pour des contaminants huileux ou gras, dégraisser la membrane avec de la laine de coton trempée dans de l'acétone ou dans une solution savonneuse. Après avoir traité l'électrode, un nouvel étalonnage doit être réalisé. Si la pente de l'électrode n'est toujours pas bonne, l'électrode doit être remplacée. 5.4 Auto-diagnostic  Lorsque l'appareil de mesure est en marche, appuyer et maintenir simultanément enfoncés les boutons Lecture et Étalonnage jusqu'à ce que l'appareil de mesure affiche l'écran complet. Chaque icône clignote l'une après l'autre. De cette manière, il est possible de vérifier si toutes les icônes sont correctement présentées. L'étape suivante consiste à vérifier que les touches fonctionnent correctement. Ceci nécessite l'interaction de l'utilisateur. Lorsque b clignote, cinq icônes sont affichées.  Appuyer sur les cinq touches dans n'importe quel ordre. Chaque fois que vous appuyez sur une touche, une icône disparaît de l'écran ; continuer d'appuyer sur les autres touches jusqu'à ce que toutes les icônes aient disparu. Lorsque l'auto-diagnostic est terminé avec succès, PAS apparaît. En cas d'échec de l'autodiagnostic, le message d'erreur Err 1 apparaît. Remarque : Vous devez terminer d'appuyer sur les cinq touches dans les 2 minutes, sinon Err 1 apparaît et vous devrez répéter la procédure. 5.5  Récupération des paramètres usine Lorsque l'appareil de mesure est à l'arrêt, maintenir simultanément appuyés les boutons Lecture, Étalonnage et Quitter pendant 3 secondes, l'écran affiche et clignote, signifie « Réinitialisation ». Deux choix se présentent à l'utilisateur :  Appuyer sur le bouton Lecture pour une réinitialisation selon les paramètres usine (MTC, pente et décalage, etc.), affichage mesure.  redémarrer ensuite l'appareil de Ou appuyer sur le bouton Quitter pour quitter le paramétrage, affichage ensuite l'appareil de mesure à l'arrêt. et mettre FR-16 STARTER 3100 6 DONNÉES TECHNIQUES 6.1 Spécifications Conditions ambiantes  Utilisation en intérieur seulement  Altitude : Jusqu'à 2000 m  Plage de température spécifiée : 5 à 40 °C  Humidité : humidité relative maximum à 80 % avec des températures jusqu'à 30 °C diminuant linéairement à 50 % d'humidité relative à 40 °C  Variations secteur jusqu'à ± 10 % de la tension nominale  Installation de catégorie II  Degré de pollution : 2  Fonctionnement assuré à température ambiante entre 5 et 40 °C Modèle ST3100 Plage de mesure pH -2,00…16,00 –1999…1999 mV -5 °C…110 °C Résolution pH 0,01 1 mV 0,1 °C Limites d'erreur pH ± 0,01 ± 1 mV ± 0,5 °C Étalonnage 3 points 3 groupes de tampons prédéfinis Mémoire 99 mesures Les dernières données d'étalonnage Alimentation Entrée Adaptateur CA : 100-240V ~ X,XA 50/60 Hz Sortie Adaptateur CA : 12V  X,XA Dimensions/poids 220 (L) x 175 (P) x 78 (H)/ 0,75 kg Affichage ACL avec rétroéclairage Entrée BNC, impédance > 10e+12 Ω Cinch, NTC 30 kΩ Entrée de référence Prise fiche banane 2 mm Compensation de température ATC et MTC Enceinte ABS STARTER 3100 7 FR-17 Conformité Ce produit est conforme à la directive EMC 2004/108/EC et à la directive de basse tension 2006/95/EC. La Déclaration de conformité est disponible en ligne sur europe.ohaus.com/europe/en/home/support/compliance.aspx. Conformément à la directive européenne 2002/96/CE sur les déchets d’équipements électriques et électroniques (WEEE), cet appareil ne doit pas être mis au rebut comme des ordures ménagères. Ceci est également valable pour les pays en dehors de l’UE, selon leurs impératifs spécifiques. Veuillez mettre au rebut ce produit conformément à la réglementation locale au point de collecte spécifié pour les équipements électriques et électroniques. En cas de questions, veuillez contacter l’autorité responsable ou le distributeur auprès duquel cet appareil a été acheté. Si cet appareil change de propriétaire (pour des raisons personnelles ou professionnelles), cette consigne doit lui être communiquée. Les instructions de mise au rebut en Europe sont disponibles en ligne sur europe.ohaus.com/europe/en/home/support/weee.aspx. Nous vous remercions de votre contribution à la protection de l’environnement. Remarque FCC : Cet équipement a été testé et déclaré conforme aux limites des appareils numériques de classe A, selon la Section 15 des règles du FCC (Commission fédérale des communications). Ces limites ont pour objectif de fournir une protection raisonnable contre des interférences dangereuses lorsque l'équipement est utilisé dans un environnement commercial. Cet équipement génère, utilise et peut rayonner des hautes fréquences et s'il n'est pas installé et utilisé conformément au guide de l'utilisateur, il peut générer des interférences préjudiciables aux communications radio. L'utilisation de cet équipement dans une zone résidentielle risque de générer des interférences préjudiciables, auquel cas l'utilisateur se verra dans l'obligation de rectifier la situation à ses frais. Enregistrement ISO 9001 En 1994, le Bureau Veritus Quality International (BVQI) a octroyé la certification d'enregistrement ISO 9001 à Ohaus Corporation, États-Unis d'Amérique, confirmant que le système de gestion de la qualité d'OHAUS était conforme aux spécifications normalisées de l'ISO 9001. Le 21 juin 2012, Ohaus Corporation, États-Unis d'Amérique, a été ré-enregistrée à la norme ISO 9001:2008. FR-18 8 STARTER 3100 Groupe de tampons Le STARTER 3100 corrige automatiquement la dépendance àla température des valeurs de pH du groupe de tampons qui sont données sur les tableaux suivants : b2, b3, vous pouvez trouver b1 dans 3.1.1) . Groupe de tampons b2 Normes européennes Temp°C pH2,00 pH4,01 pH7,00 pH9,21 5 2,02 4,01 7,09 9,45 10 2,01 4,00 7,06 9,38 15 2,00 4,00 7,04 9,32 20 2,00 4,00 7,02 9,26 25 2,00 4,01 7,00 9,21 30 1,99 4,01 6,99 9,16 35 1,99 4,02 6,98 9,11 40 1,98 4,03 6,97 9,06 45 1,98 4,04 6,97 9,03 50 1,98 4,06 6,97 8,99 pH11,00 11,72 11,54 11,36 11,18 11,00 10,82 10,64 10,46 10,28 10,10 Groupe de tampons b3 JJG119 Temp °C pH1,68 pH4,00 pH6,86 5 1,67 4,00 6,95 10 1,67 4,00 6,92 15 1,67 4,00 6,90 20 1,68 4,00 6,88 25 1,68 4,00 6,86 30 1,68 4,01 6,85 35 1,69 4,02 6,84 40 1,69 4,03 6,84 45 1,70 4,04 6,83 50 1,71 4,06 6,83 pH12,46 13,21 13,01 12,82 12,64 12,46 12,29 12,13 11,98 11,83 11,70 pH9,18 9,39 9,33 9,28 9,23 9,18 9,14 9,11 9,07 9,04 9,02 STARTER 3100 FR-19 GARANTIE LIMITÉE Ohaus garantit que ses produits sont exempts de défauts matériels et de fabrication à compter de la date de livraison pendant toute la durée de la garantie. Pendant la période de garantie, Ohaus réparera ou, à sa convenance, remplacera sans frais toutes les pièces déterminées défectueuses, sous réserve que le produit soit retourné, fret payé d'avance, à Ohaus. Cette garantie ne s'applique pas si le produit a subi des dommages suite à un accident ou un mésusage, a été exposé à des matériaux radioactifs ou corrosifs, contient des matériaux étrangers ayant pénétré à l'intérieur ou après une maintenance ou des modifications apportées par quiconque autre qu’Ohaus. En l’absence d’une carte d'enregistrement de garantie dûment remplie, la période de garantie commence à la date de l'expédition au revendeur agréé. Aucune autre garantie expresse ou implicite n'est offerte par Ohaus Corporation. En aucun cas, Ohaus Corporation ne peut être tenu responsable des dommages indirects. Dans la mesure où les lois régissant les garanties varient d'un État à un autre et d'un pays à un autre, veuillez contacter Ohaus ou votre représentant local agréé Ohaus pour de plus amples informations. 1 Índice Introdução ............................................................................1 1.1 1.2 1.3 2 Instalação .............................................................................5 2.1 2.2 2.3 2.4 2.5 2.6 3 Conteúdo da embalagem ............................................................... 5 Instalando o suporte de eletrodo independente .......................... 6 Instalando o adaptador de força .................................................... 6 Conecte os eletrodos de pH........................................................... 7 Suporte do medidor para ajuste do ângulo de visão .................. 7 Guia rápido incluso ......................................................................... 7 Configuração........................................................................8 3.1 3.2 3.3 4 Definição de sinais e símbolos de aviso ...................................... 1 Precauções de segurança .............................................................. 2 Display e controles ......................................................................... 3 Configurar a unidade de temperatura ........................................... 8 Configure a compensação manual de temperatura .................... 8 Selecionando um grupo predefinido de buffer ............................ 8 Funcionamento do STARTER 3100 ....................................9 4.1 Calibração ........................................................................................ 9 4.1.1 Grupo de buffer ............................................................................. 9 4.1.2 Executando calibração de 1 ponto .............................................. 10 4.1.3 Executando calibração de 2 pontos .............................................11 4.2 Medição de amostra .......................................................................11 4.2.1 Medição de pH .............................................................................11 4.2.2 Medição de mV ............................................................................11 4.3 Medição de temperatura ................................................................11 4.4 Utilizando a memória .................................................................... 12 4.4.1 Armazenando uma leitura ........................................................... 12 4.4.2 Acessando a memória ................................................................ 12 4.4.3 Apagando a memória .................................................................. 12 4.5 Impressão ...................................................................................... 13 5 Manutenção ........................................................................ 14 5.1 5.2 5.3 5.4 5.5 6 7 Mensagens de erro........................................................................ 14 Manutenção do medidor ............................................................... 14 Manutenção do eletrodo ............................................................... 14 Autodiagnóstico ............................................................................ 15 Recuperando as configurações de fábrica ................................ 15 Especificações ................................................................... 16 Grupo de buffers................................................................ 17 STARTER 3100 1 PT-1 Introdução Grato por optar por um produto da OHAUS. Favor ler completamente o manual antes de utilizar o medidor de pH de bancada STARTER 3100 para evitar funcionamento incorreto. O STARTER 3100 possui excelente relação preço/performance, sendo projetado com muitos recursos úteis, incluindo um suporte de eletrodo independente, um bepper para alerta de erro. Também oferecemos outros acessórios, por exemplo, eletrodo de pH para diferentes aplicações e soluções buffer. Você achará muitas funcionalidades úteis nesses medidores. Algumas delas são:  LCD grande com iluminação com guia rápido incluso, de fácil operação pelo usuário  Ponto final automático/manual que pode auxiliar os usuários a congelar o valor de leitura estável  Três grupos de buffer (US, EU e JJG) atendem a diferentes requisitos  O RS232 pode conectar uma impressora para impressão, 99 memórias de dados com os dados da última calibração 1.1 Definição de sinais e símbolos de aviso Notas de segurança são marcadas com palavras e símbolos de aviso. Estas mostram avisos e questões de segurança. Ignorar as notas de segurança pode levar a lesões pessoais, danos ao instrumento, avarias e resultados falsos. Palavras de sinais AVISO Para uma situação perigosa com risco médio, possivelmente resultando em ferimentos ou morte se não for evitada. CUIDADO Para uma situação perigosa com baixo risco, resultando em danos ao aparelho ou a propriedade ou perda de dados, ou lesões, se não for evitada. Atenção Para uma informação importante a respeito do produto. Nota Para obter informações úteis sobre o produto Símbolos de aviso Risco geral Risco de explosão Risco de corrosão  Corrente alternativa PT-2  1.2 STARTER 3100 Corrente direta Precauções de segurança AVISO: Leia todos os avisos de segurança antes de instalar, fazer conexões ou manutenção deste equipamento. Não seguir estes avisos pode resultar em lesões pessoais ou danos à propriedade. Reter todas as instruções para futura referência.           Verifique que a faixa de tensão de entrada impressa na etiqueta de dados e o tipo de plugue de tomada estão de acordo com os padrões locais de adaptação. Certifique-se de que o cabo de alimentação não represente um obstáculo em potencial ou o perigo de tropeçar. Utilize o equipamento apenas em locais secos. Seque qualquer líquido que derramar imediatamente. O instrumento não é à prova d'água. Quando utilizar produtos químicos e solventes, siga as instruções do produtor químico e as regras de segurança geral do laboratório. Utilize somente acessórios aprovados. Opere o equipamento somente em condições ambientais especificadas neste manual. Desconecte o equipamento da fonte de alimentação para limpeza. Não opere o equipamento em ambientes perigosos ou instáveis. Serviço deve ser realizado apenas por pessoal autorizado. STARTER 3100 1.3 PT-3 Display e controles Displays 1 2 3 4 5 6 7 1 8 9 Condição do eletrodo Slope: mais de 95% e offset: ± (0-15) mV Condição do eletrodo é boa Slope: 90-95% ou offset: ± (15-35) mV Condição do eletrodo é aceitável Slope: menos que 90% ou offset: ± (35-60) mV Condição do eletrodo não é boa e necessita limpeza 2 Ícone de estabilidade de ponto final 3 ; ícone de ponto final automático Ícone de medição - 4 Ícone de calibração - 5 Ícone de configuração ; instrumento está no modo de configuração, pode configurar temperatura (MTC), grupo de buffer, etc. 6 Leitura pH/mV ou slope em processo de calibração 7 Ponto de calibração 8 Compensação automática de temperatura - ATC ; Compensação manual de temperatura MTC 9 Temperatura durante medição ou offset (mV) em processo de calibração ; calibração ou medição em andamento ; calibração em andamento / Grupo de Buffer / Número de memória / Índice de erro PT-4 STARTER 3100 Controles Botão Pressione e solte Pressione e mantenha pressionado por 3 segundos - Iniciar ou terminar a medição - Confirmar configuração, armazenar valor inserido - Ativar/desativar ponto final automático - Iniciar a calibração - Acessar os últimos dados de calibração: slope e offset - Ativar o medidor - Sair e retornar à tela de medição - Desativar o medidor - Armazenar leitura atual na memória - Aumentar valor durante ajuste - Rolar para cima na memória - Acessar dados armazenados - Imprimir dados da memória atual - Alternar entre pH e mV - Diminuir valor durante ajuste - Rolar para baixo na memória - Entrar no modo de configuração , - Iniciar autodiagnóstico Ativar/desativar a iluminação do LCD STARTER 3100 2 PT-5 Instalação Com cuidado, abra a caixa. 2.1 Conteúdo da embalagem O modelo ST3100-B (embalagem básica) possui os seguintes itens: ST3100-B Unidades STARTER 3100 1 Porta eletrodo independente 1 unidade Fonte de força 12V 1 unidade Além do conteúdo ST3100-B, a embalagem ST3100-F também incluiu o seguinte: Conjunto de Pó Buffer para pH (4.01, 7.00, 10.01) 1 unidade Eletrodo de pH recarregável 3 em 1 ST310 1 Cobertura de proteção 1 Além do conteúdo ST3100-B, a embalagem ST3100-H também incluiu o seguinte: Conjunto de Pó Buffer para pH (4.01, 7.00, 10.01) 1 unidade Eletrodo de pH gel 3 em 1 ST320 1 Cobertura de proteção 1 Cada pó buffer de pH deverá ser dissolvido em 250ml de água pura ou água deionizada em um frasco volumétrico. Você também pode pedir da Ohaus solução buffer em frasco. Acessórios: Modelo Descrição Item NO Eletrodo de pH recarregável de plástico 3 em 1 ST310 83033965 Eletrodo de pH recarregável de plástico 2 em 1 ST210 83033966 Eletrodo de pH com gel plástico 3 em 1 (sem ST320 necessidade de recarregar) 83033967 Eletrodo de pH de amostra lamacenta de vidro 2 ST230 em 1 83033968 Eletrodo de pH de amostra de água pura de STPURE vidro 2 em 1 83033969 STTEMP30 Eletrodo de Temperatura 83033970 STORP1 Eletrodo ORP de plástico gel 30038555 STORP2 Eletrodo ORP de vidro recarregável 30038553 STREF01 Eletrodo de referência Ag/AgCl 30059253 PT-6 STARTER 3100 Conjunto de pó de buffer (4.01; 7.00; 10.0) Buffer pH4.01 250ml Buffer pH7.00 250ml Buffer pH9.21 250ml Buffer pH10.00 250ml Eletrólito de Referência de eletrodo de pH Solução de proteção de eletrodo de pH Frasco de proteção de sensor de pH (10 na embalagem) Suporte de eletrodo independente (3100/3100C) Impressora SF40A Cobertura de proteção 2.2 83033971 30065083 30065084 30065085 30065086 30059255 30059256 30064800 30058733 30045641 30058734 Instalando o suporte de eletrodo independente Instale o braço do eletrodo na base, Afrouxar Apertar Após a ajustar o botão de tensão em certa medida, você poderá levantar e abaixar diretamente o braço. Quando colocamos um eletrodo de pH no braço e executamos calibração ou medição, podemos colocar o frasco de armazenamento do eletrodo de pH no orifício da base. 2.3 Instalando o adaptador de força Insira o clipe adaptador correto na ranhura do adaptador de força. A fonte de força da balança Ohaus também é 12V, podendo também ser utilizada para o medidor de pH. STARTER 3100 2.4 PT-7 Conecte os eletrodos de pH Há 3 soquetes para o eletrodo. Soquete "pH" (BNC); Soquete “TEMP.” (Cinch) e Soquete “REF” (banana 2 mm). Para o eletrodo ST310, deve-se conectar o soquete BNC e Cinch. Para eletrodo pH 2 em 1 (ST210), só é necessário conectar o BNC. Soquete REF para eletrodo de referência separado. (por exemplo, STREF01) 2.5 Suporte do medidor para ajuste do ângulo de visão Um recurso exclusivo do STARTER 3100 é o suporte do medidor, que pode ser utilizado para ajustar o ângulo de visão do display caso trabalhe em uma mesa alta. 2.6 Guia rápido incluso Um outro recurso exclusivo do STARTER 3100 é o guia rápido incluso, que está afixado no gabinete inferior do medidor, podendo utilizar o dedo para pressioná-lo para sair. PT-8 STARTER 3100 3 Configuração 3.1 Configurar a unidade de temperatura Por favor, note: Se for utilizado um eletrodo de temperatura, compensação automática de temperatura (ATC) e a temperatura da amostra são exibidos na tela. Em seguida, você pode optar por ignorar a configuração MTC (abaixo). Se o medidor não detecta um eletrodo de temperatura ou um não é usado, o medidor automaticamente alterna para o modo de compensação de temperatura Manual (MTC) e MTC aparece na tela. MTC pode ser definido como segue:  Ligar o medidor pressionando Exit.  Pressionar e manter o botão Mode/Setup pressionado até o ícone de setup aparecer no display e a unidade de temperatura atual piscar.  Pressionar o botão up ou down para intercalar entre °C e °F .  Pressionar o botão Read para confirmar sua seleção. A seguir  Continue com a configuração da temperatura MTC ou  Pressione Read para confirmar a configuração  Pressione o botão Exit para retornar à tela de medição. 3.2 Configure a compensação manual de temperatura Após confirmar a unidade de temperatura, aparecerá no display o valor de temperatura MTC (Compensação Manual de Temperatura). A seguir, utilize o botão up ou down para aumentar ou diminuir o valor até a temperatura de sua amostra. Pressione o botão Read para confirmar a configuração ou pressione o botão Exit para retornar à tela de medição. O valor padrão da temperatura MTC é 25 °C (77 °F). Nota: °C = 5/9 (°F -32) 3.3 Selecionando um grupo predefinido de buffer Após confirmar o valor da temperatura de compensação MTC, acessa-se a seleção de grupo de buffer. Utilize botão up ou down para selecionar um grupo de buffer entre os quatro grupos existentes (3.1.1). Pressione o botão Read para confirmar a configuração ou o botão Exit para sair e retornar à tela de medição. O grupo-padrão de buffer é b1. b1 1.68 4.01 7.00 10.01 STARTER 3100 4 PT-9 Funcionamento do STARTER 3100 O procedimento de medição de pH é o seguinte; os pontos principais são o eletrodo de pH, enxágue, calibração, armazenamento. a) preparação do eletrodo de pH b) preparação do buffer e calibração do eletrodo de pH c) preparação da amostra d) preparação do pH e) registro ou impressão do resultado da medição f) enxágue e armazenamento do eletrodo de pH Preparação do eletrodo de pH: o eletrodo de pH deverá ser de enxaguado com água pura antes e após a utilização. Verifique se o eletrodo está fisicamente danificado (tome cuidado com o bulbo de vidro) O eletrodo de pH deverá ser armazenado no frasco de armazenamento; a solução no frasco é uma solução 3M KCl. Após o eletrodo de pH ser colocado na amostra ou solução buffer, o usuário deverá agitar vários segundos e, a seguir, aguardar 30 a 60 segundos para que o sinal se estabilize e, em seguida, pressionar o botão para funcionar (Calibração ou medição). 4.1 Calibração 4.1.1 Grupo de buffer STARTER 3100 O STARTER 3100 pode executar calibrações de 1, 2 ou 3 pontos. Há 3 grupos de buffer no medidor; você poderá selecionar o grupo de buffer que preferir (ver 4.3); o buffer padrão é b1 US standards; o valor de buffer será automaticamente reconhecido durante a calibração. Os 3 grupos de buffer predefinidos são (a 25 °C): b1 1.68 4.01 7.00 10.01 b2 2.00 4.01 7.00 9.21 11.00 b3 1.68 4.00 6.86 9.18 12.46 O STARTER 3100 automaticamente corrige a dependência de temperatura dos valores de pH de buffer estabelecidos na seguinte tabela – grupo de buffer b1. 5 °C 1.67 4.01 7.09 10.25 10 °C 1.67 4.00 7.06 10.18 15 °C 1.67 4.00 7.04 10.12 20 °C 1.68 4.00 7.02 10.06 25 °C 1.68 4.01 7.00 10.01 30 °C 1.68 4.01 6.99 9.97 35 °C 1.69 4.02 6.98 9.93 40 °C 1.69 4.03 6.97 9.89 45 °C 1.70 4.05 6.97 9.86 50 °C 1.71 4.06 6.96 9.83 PT-10 STARTER 3100 Calibração: o eletrodo de pH precisa ser calibrado com solução buffer padrão de pH antes de uma medição correta de pH. A calibração exibirá o valor correto de pH quando o medidor receber o sinal do valor mV do eletrodo de pH. Slope: coeficiente linear entre mV e pH de acordo com o valor teórico (por exemplo, 59.16mV/pH @ 25 °C, que significa 100% slope); Offset: o valor mV quando o valor de pH for 7.00. (Valor teórico é 0 mV); 4.1.2 Executando calibração de 1 ponto Ao realizar a calibração, Ohaus recomenda o uso do modo de ponto de extremidade do Auto. De pois de ligar otaxímetro, certifiquese o topo da tela mostra para garantir que o medidor está no modo de ponto final de Auto. Modo Endpoint: Temos dois modos endpoint - auto endpoint e manual endpoint. Ao pressionar e manter pressionado o botão Read, você poderá alternar entre os 2 modos de ponto final. o Para executar o ponto final manualmente de uma medição ou calibração, pressione o botão Read: a leitura da amostra congela; o ícone de estabilidade de ponto final pisca 3 vezes e congela no display. o Quando estiver no modo de ponto final automático, o medidor considerará se a leitura está estável e, a seguir, executará automaticamente o ponto final; a leitura congela e pisca 3 vezes e, em seguida, desaparece; os ícones de ponto final automático piscam 3 vezes e congelam no display. o Critério de estabilidade: o sinal de entrada do sensor não pode mudar em mais de 0.1mV em 6 segundos. Nota: Com a calibração de 1 ponto, somente o offset é ajustado. Se o sensor foi anteriormente calibrado com calibração multiponto, o slope anteriormente armazenado permanecerá. Caso contrário, será utilizado o slope 100 % teórico (-59.16 mV / pH). Quando o STARTER 3100 estiver no modo de medição de pH (ver 3.2), coloque o eletrodo de pH em um buffer de calibração e a seguir:  Pressione o botão –Cal; “Cal 1” é exibido na parte inferior esquerda da tela e “Cal ” pisca. e aparecem na parte superior da tela, pisca durante a calibração.  O medidor alcança o ponto final de acordo com o modo endpoint, o valor de pH do ponto de calibração (por exemplo, 7.00) com a exibição de temperatura na tela. A calibração de 1 ponto terminou; agora, temos 3 escolhas a seguir:  Pressionar o botão Cal para executar a calibração de 2 pontos.  Pressionar o botão Read para armazenar a calibração de 1 ponto e sair; o offset e o slope são mostrados no display por 3 segundos; a seguir, retorna para a tela de medição.  Pressionar o botão Exit para rejeitar a calibração e voltar à tela de medição. STARTER 3100 4.1.3     PT-11 Executando calibração de 2 pontos Execute a calibração de 1 ponto conforme descrita acima. Enxágue o eletrodo de pH com água pura e enxugue com o papel toalha. Coloque o eletrodo no próximo buffer de calibração e pressione o botão Cal, “Cal 2” será exibido no canto esquerdo inferior da tela e “Cal ” piscará na parte superior da tela. e aparecerão, piscará durante a calibração. O medidor alcança o ponto final de acordo com o modo de ponto final, o valor de pH do ponto de calibração (por exemplo, 4.01) com a exibição da temperatura na tela. A calibração de 2 pontos está finalizada; agora temos 3 escolhas a seguir:  Pressionar o botão Cal para executar a calibração de 3 pontos.  Pressionar o botão Read para armazenar a calibração de 2 pontos e sair; são exibidos no display o offset e o slope por 3 segundos e, a seguir, retorna à tela de medição.  Pressionar o botão Exit para rejeitar a calibração e retornar à tela de medição. Nota: A utilização de um eletrodo de temperatura ou um eletrodo com o sensor de temperatura integrado (eletrodo de pH 3 em 1) é recomendada. Se utilizar o modo MTC, deve inserir o valor correto da temperatura e manter todas as soluções de amostra e buffer na temperatura ajustada. (ver 4.2) 4.2 Medição de amostra 4.2.1 Medição de pH    Coloque o eletrodo de pH na amostra Pressione o botão Read para iniciar a medição de pH; aparecerá no display. piscará durante a medição. Quando o medidor alcançar o ponto final, o valor de pH com a exibição de temperatura serão mostrados na tela. 4.2.2   4.3 Medição de mV Pressione o botão Mode para intercalar entre o modo de medição de pH e o modo de medição de mV. Siga o mesmo procedimento para medição de pH para executar a medição de mV. Medição de temperatura Para melhor precisão, recomendamos utilizar um eletrodo de temperatura integrado ou um eletrodo de temperatura separado.  Se um eletrodo de temperatura for utilizado, ATC e a temperatura da amostra serão exibidos.  Se o medidor não detectar um eletrodo temperatura, passará automaticamente para o modo de compensação manual de temperatura e aparecerá MTC A temperatura MTC deverá ser ajustada. PT-12 STARTER 3100 Nota: O STARTER 3100 aceita sensor de temperatura NTC 30 kΩ. ATC (Compensação Automática de Temperatura) ou MTC (Compensação Manual de Temperatura), a Compensação de Temperatura somente corrige alteração na saída do eletrodo, não alteração na solução real. Isso significa que o medidor corrige o sinal (mV) dos eletrodos de pH para garantir um valor de pH mais preciso de acordo com a temperatura real. 4.4 Utilizando a memória 4.4.1 Armazenando uma leitura O STARTER 3100 pode armazenar até 30 resultados de pontos finais.  Pressione o botão Store quando a medição alcançar o ponto final. M01 indica que um resultado foi armazenado. Se pressionar o botão Store quando M99 for exibido, FUL será exibido, indicando que a memória está cheia. Para armazenar mais dados, você terá que liberar espaço na memória. (Ver 3.4.3) 4.4.2    Acessando a memória Pressione e mantenha o botão Recall pressionado para acessar os valores armazenados na memória quando a medição atual alcançar o ponto final. Pressione o botão up ou down para rolar pelos resultados armazenados. R01 a R30 indicam que o resultado está sendo exibido. Pressione o botão Exit para sair. 4.4.3 Apagando a memória Pressione o botão up ou down para navegar pelos resultados armazenados até “MRCL” aparecer.  Pressione o botão Read, CLr piscará; Agora, temos duas escolhas:  Pressione o botão Read para confirmar o apagamento de todos os dados armazenados.  Pressione o botão Exit para retornar ao modo de medição sem apagar a memória.  STARTER 3100 4.5 PT-13 Impressão Se a impressora estiver conectada ao STARTER 3100 (por exemplo, SF-F40A, a luz verde significa sucesso da conexão), será gerada automaticamente uma impressão após cada calibração ou medição com ponto final. O formato da impressão após uma medição de pH será: Ponto Final, Valor, Temperatura, ATC/MTC Auto EP, 4.01pH, 25.0 °C, MTC Os detalhes da segunda linha são: Auto EP, 4.01pH, 25.0 °C , MTC | | | |---- Compensação Manual de Temperatura | | |---------------- Valor de temperatura e unidade °C | |------------------------- Valor de pH |---------------------------------------- Ponto Final Automático O formato da impressão após uma medição de mV é: Ponto Final, Valor, Temperatura, ATC/MTC Manual EP, 182mV, 23.2 °C, ATC A impressão em caso de erro de mensagem será: Ponto Final, Valor, Temperatura, ATC/MTC Error3 A impressão para uma calibração de 2 pontos será: Buffer1: 4.01pH mV1: 178mV Temp.1: 25.0 °C Buffer2: 7.00pH mV2: 3mV Temp.2: 25.0 °C Slope: 99% Offset: 5mV ATC/MTC: MTC A impressão para uma calibração de 3 pontos será: Buffer1: 4.01pH mV1: 178mV Temp.1: 25.0 °C Buffer2: 7.00pH mV2: 0mV Temp.2: 25.0 °C Buffer3: 9.21pH mV3: -130mV Temp.3: 25.0 °C Slope: 100% Offset: 0mV ATC/MTC: ATC Quando navegar pela memória, você poderá imprimir a entrada atualmente exibida pressionando e mantendo pressionado (3 segundos) o botão Store. O formato da impressão será o seguinte: Recall 01: Ponto Final, Valor, Temperatura, ATC/MTC Manual EP, 4.01pH, 35.6 °C, ATC PT-14 STARTER 3100 5 Manutenção 5.1 Mensagens de erro Error 0 Erro de acesso à memória Volta para as configurações de fábrica Error 1 Falha de autodiagnóstico Repita o procedimento de autodiagnóstico e verifique se terminou de pressionar todas as cinco teclas dentro de 2 minutos. Error 2 Valores medidos fora de faixa Verifique se o eletrodo está adequadamente conectado e colocado na solução de amostra. Error 3 Temperatura medida de buffer fora da faixa (<5 ou >40 °C) Offset fora da faixa offset > 60mV ou < - 60 mV Mantenha a temperatura de buffer de pH dentro da faixa de calibração Error 5 Slope fora da faixa Verifique se o buffer de pH está correto e é novo; Limpe ou substitua o eletrodo de pH. Error 6 Medidor não reconhece o buffer Verifique se o buffer é correto e é novo; verifique se o buffer não foi utilizado mais do que uma vez. Error 9 Conjunto de dados atuais já foi armazenado antes Error 10 Temperatura de amostra fora da faixa Uma leitura de ponto final somente pode ser armazenada uma vez. Execute uma nova medição para armazenar. Verifique a temperatura de amostra, o sensor de temperatura. Error 4 Verifique se o buffer de pH está correto e é novo; Limpe ou substitua o eletrodo de pH. Se o erro ocorrer, o medidor também irá bipar três vezes para alerta. Quaisquer outros problemas, contate o representante de vendas da OHAUS ou envie um email para [email protected]. 5.2 Manutenção do medidor Jamais desparafuse as duas metades do gabinete! Os instrumentos da série STARTER 3100 não requerem nenhuma manutenção, exceto limpeza ocasional com tecido umedecido e a substituição das baterias esgotadas. O gabinete é feito de acrilonitrilo butadieno estireno (ABS). Esse material é atacado por alguns solventes orgânicos, por exemplo, tolueno, xileno e metiletilcetona (MEK). Qualquer respingo deverá ser imediatamente limpo. 5.3 Manutenção do eletrodo Verifique se o eletrodo está cheio de solução de eletrólito. Sempre armazene o eletrodo de acordo com o seu manual de instrução e não deixe secar. STARTER 3100 PT-15 Se a resposta do eletrodo tornar-se lenta ou se o slope não for suficientemente bom, tente: o Deixar o eletrodo em imerso 0.1M HCl por mais de 8 horas. o Para contaminante óleo ou gordura, desengordure a membrana com algodão saturado em acetona ou solução saponácea. Após o tratamento do eletrodo, deverá ser executada uma nova calibração. Se o slope do eletrodo ainda não for bom, pode ser necessário substituir o eletrodo. 5.4 Autodiagnóstico  Quando o eletrodo estiver ativado, pressione e mantenha pressionado o botão Read e o botão Cal simultaneamente até que o eletrodo exiba a tela cheia. Cada ícone piscará um após o outro. Dessa forma, você poderá verificar se todos os ícones são corretamente exibidos. A próxima etapa é verificar se as teclas estão funcionando corretamente. Isso exige interação do usuário. Quando b piscar, cinco ícones serão exibidos.  Pressione as cinco teclas em qualquer ordem. Cada vez que pressionar uma tecla, um ícone desaparecerá da tela; continue pressionando as outras teclas até que todos os ícones tenham desaparecido. Quando terminar o autodiagnóstico com sucesso, aparecerá PAS. Se houver falha no autodiagnóstico, a mensagem de erro Err 1 será exibida. Nota: Você terá que terminar de pressionar todas as cinco teclas dentro de 2 minutos; caso contrário, aparecerá a mensagem Err 1 e você terá que repetir o procedimento. 5.5  Recuperando as configurações de fábrica Quando o medidor estiver desligado, pressione e mantenha pressionados os botões Read, Cal e Exit em conjunto por 3 segundos; a tela exibirá e piscará, significando “Reset”. A seguir, temos 2 escolhas:  Pressione o botão Read para voltar às configurações de fábrica (MTC, slope e offset, etc.), exibirá  ; a seguir, reinicie o medidor. Ou pressione o botão Exit para sair da configuração; exibirá o medidor. ; a seguir, desligue PT-16 6 STARTER 3100 Especificações STARTER 3100 Faixa de medição -2.00…16.00 pH –1999…1999 mV -5 °C…110 °C Resolução 0.01 pH 1 mV 0.1 °C Limites de erro ± 0.01 pH ± 1 mV ± 0.5 °C Calibração 3 pontos 3 grupos predefinidos de buffer Memória 99 medições Últimos dados de calibração Alimentação 110-240V/50Hz, DC 12V Tamanho/peso Aproximadamente 220 L x 175 D x 78 A mm / 0.75 kg Display Cristal líquido com iluminação Entrada BNC, impedância > 10e+12 Ω Cinch, NTC 30 kΩ Entrada de referência Soquete banana de 2 mm Compensação de temperatura ATC & MTC Gabinete ABS STARTER 3100 7 PT-17 Grupo de buffers O STARTER 3100 corrige automaticamente a dependência de temperatura do valor de pH do grupo buffer estabelecido nas tabelas a seguir (b2, b3); você pode encontrar b1. (ver 3.1.1) Grupo de buffer b2 Temp °C pH2.00 5 2.02 10 2.01 15 2.00 20 2.00 25 2.00 30 1.99 35 1.99 40 1.98 45 1.98 50 1.98 Padrão Europa pH4.01 pH7.00 4.01 7.09 4.00 7.06 4.00 7.04 4.00 7.02 4.01 7.00 4.01 6.99 4.02 6.98 4.03 6.97 4.04 6.97 4.06 6.97 pH9.21 9.45 9.38 9.32 9.26 9.21 9.16 9.11 9.06 9.03 8.99 pH11.00 11.72 11.54 11.36 11.18 11.00 10.82 10.64 10.46 10.28 10.10 Grupo de buffer b3 Temp °C pH1.68 5 1.67 10 1.67 15 1.67 20 1.68 25 1.68 30 1.68 35 1.69 40 1.69 45 1.70 50 1.71 JJG119 pH4.00 4.00 4.00 4.00 4.00 4.00 4.01 4.02 4.03 4.04 4.06 pH9.18 9.39 9.33 9.28 9.23 9.18 9.14 9.11 9.07 9.04 9.02 pH12.46 13.21 13.01 12.82 12.64 12.46 12.29 12.13 11.98 11.83 11.70 pH6.86 6.95 6.92 6.90 6.88 6.86 6.85 6.84 6.84 6.83 6.83 STARTER 3100 Bench pH Meter EN-19 Ohaus Corporation 7 Campus Drive Suite 310 Parsippany, NJ 07054 USA Tel: +1 (973) 377-9000 Fax: +1 (973) 944-7177 With offices worldwide/Con oficinas en todo el mundo/Avec des bureaux dans le monde entier/Com escritórios no mundo inteiro www.ohaus.com *30031621* PN 30031621 E © Ohaus Corporation 2016, all rights reserved/todos los derechos reservados/tous droits réservés/todos os direitos reservados
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88

Ohaus ST3100-B Manual de usuario

Categoría
Medir, probar
Tipo
Manual de usuario
Este manual también es adecuado para