Eutech CON El manual del propietario

Categoría
Medición
Tipo
El manual del propietario
2. When the READY indicator appears use up/down arrows to
adjust the primary reading to match the standard value at
the measured temperature, then press ENTER.
3. Rinse your electrode(s) then dip the sensor into your next
standard, or press MEAS to return to measurement mode.
Repeat as necessary. After the desired number of standards
have been calibrated, the bottom display will show the
efficiency in measurement mode.
Temperature Calibration/Manual ATC
Temperature calibration is recommended prior to first use,
after ATC sensor replacement, and periodically as needed.
1. Press CAL from any measurement, then press MODE.
2. Skip to step 3 for manual ATC, otherwise, dip the temperature
sensor into a solution with a known accurate temperature.
The upper display shows the active temperature while
the lower display shows the factory default temperature
without adjustment.
3. Use up/down arrows to adjust the upper display.
Press ENTER to accept the calibration temperature.
The maximum adjustable value is ±10 ºC (or ±18 ºF) from
factory default.
Error Messages
“ERR” will appear when an error condition exists or the
incorrect key is pressed. Common examples include:
• Pressing ENTER during calibration before the READY
indicator appears. Wait for the READY indicator before
pressing ENTER.
UR (Under Range) • OR (Over Range)
Intended Use, Maintenance & Precautions
These handheld meters use sensors to detect various parameters
for water-based measurements. For routine maintenance
disconnect the power cord or battery, then dust or wipe the
display using a damp cloth. If necessary, warm water or a mild
water based detergent can be used. Immediately remove any
spilled substance from contact with the meter using the proper
cleaning procedure for the type of spill.
Do not use this equipment in potentially explosive atmospheres.
Refer to the electrode instructions for use, storage and cleaning.
Ensure that no liquid enters the instrument.
Do not use any aggressive cleaning chemicals (solvents or
similar agents).
• There are no user serviceable parts inside. Attempts to
service internal parts may void the warranty.
WARNING: No modification of this equipment is allowed.
Total Dissolved Solids Factor. Adjust from (0.00 - 1.00)
Normalization Temperature. Choose (15.0 / 20.0 / 25.0 /
30.0 °C)
• Choose Single-Point Calibration (SPC) to apply a single
calibration value across all ranges or Multi-Point Calibration
(MPC) to calibrate individual range(s).
Choose Automatic Calibration Standards or Manual Adjustment
Select Calibration Due Reminder
Set number of days from 0-60 for desired parameter
View Calibration Data (for the parameter being measured)
• Press ENTER to view each point that is calibrated with the
associated calibration range.
View Electrode Data (for the parameter being measured)
• Press ENTER to view the electrode efficiency of each point
that is calibrated with the associated calibration range.
System Settings
Data Logging:
MANUAL upon key press only
TIMED interval. Choose (SEC / MIN / HOUR) interval.
Automatic shut off after 10 minutes. Choose ON or OFF.
Clock Settings:
Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY).
Time: Choose (24HR or 12HR). If 12HR, choose AM or PM.
Set Printer Type:
CSV (Comma Separated Values) – best format for computer
Printer (Text) – best format for printer.
Choose Manual (MAN) upon key press or TIMED interval.
If timed, choose (SEC / MIN / HOUR).
Reset
NO. Exits from reset menu options without action.
FACTORY RESET. Returns all settings except date/time
and ATC calibration to factory default values after ENTER is
pressed then restarts meter.
DATA RESET. Erases data stored in memory while retaining
other settings after ENTER is pressed.
CALIBRATION RESET. Erases non-ATC calibration data while
retaining other settings after ENTER is pressed.
Conductivity/TDS/Salinity Calibration
For best results, periodic calibration with known accurate
standards is recommended. Calibrate with standard(s) near
your intended measuring range. Provide stirring for best results.
After calibration, the electrode efficiency that corresponds to
the active measurement will be visible on the bottom display.
“- - -“ will be shown if no calibration is performed. Press MEAS
to return to measurement mode at any time.
Using Automatic Conductivity Recognition
1. From Conductivity measurement mode, dip the sensor
into your standard – either 84 μS, 1413 μS, 12.88 mS,
or 111.8 mS, then press CAL. The primary display will search
for the nearest standard. The secondary display will show
the un-adjusted value and the bottom display will show the
temperature. One calibration value in each range is allowed.
2. When the READY indicator appears, press ENTER to accept.
The primary reading will flash DONE”.
3. To calibrate another standard, rinse your electrode(s) then
dip into your the next standard or press MEAS to return
to measurement mode at any time. The primary display
will search for the nearest standard value that has not yet
been calibrated, while the secondary display will show
the unadjusted value. When the READY indicator appears,
press ENTER to accept.
4. After the desired number of standards has been calibrated,
the bottom display will show the efficiency in measurement
mode.
Using Manual Adjustment
1. From Conductivity/TDS/SALINITY measurement mode, dip
the sensor into your standard then press CAL.
150 & 450 Series Waterproof Handheld Meter
Conductivity/TDS/Salinity Operation Instructions
150 & 450 Series Medidor de mano resistente al agua
Conductividad/TDS/Salinidad Instrucciones de funcionamiento
Part of Thermo Fisher Scientific Parte de Thermo Fisher Scientific
CON 150
Con/TDS
CON 450
Con/TDS/Salinity
PC 450
pH/mV/Con/TDS
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-16/24, Singapore 139949
Tel: (65) 6778-6876
Fax: (65) 6778-0086
eutech@thermofisher.com
www.eutechinst.com
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-16/24, Singapore 139949
Tel: (65) 6778-6876
Fax: (65) 6778-0086
eutech@thermofisher.com
www.eutechinst.com
Press once to power ON in the mode that was
previously used. Press again to turn backlight on
for one minute or off (450 series only). Hold for 3
seconds to power OFF.
Toggle between measurement and calibration
modes. In SETUP mode, BACK serves to return to
the previous menu option or setting.
Confirm calibration values in CAL mode.
Confirm selections in SETUP mode.
Freeze or release the measured reading.
Customize instrument settings and preferences.
(See also Setup Programs)
Toggle between available measurement types.
Save measurement into memory. Increase value
or scroll up in SETUP or manual calibration.
Recall saved values from memory. Decrease value
or scroll down in SETUP or manual calibration.
Send output data to printer or computer.
(450 series only).
Range
#
Conductivity
Range
Automatic Calibration Values
Normalization Temperature
25 ºC 20 ºC
r 1 0.00 - 19.99 µS None None
r 2 20.0 - 199.9 µS 84 µS 76 µS
r 3 200.0 - 1999 µS 1413 µS 1278 µS
r 4 2.00 - 19.99 mS 12.88 mS 11.67 mS
r 5 20.0 - 200.0 mS 111.8 mS 102.1 mS
Models:
Getting Started/Connections
After installing (2) AA batteries and/or connecting the optional
110/220 VAC power supply, connect the desired sensors to the
corresponding ports.
12 mm and 16 mm probes can utilize the Grip-Clip™ to attach
one or more sensors to a beaker and to the instrument as
needed. The stand can be extended as shown above or used
for wall-mounting.
Keypad Functions
Setup Programs
To access the settings below, press SETUP. Up/down arrows
will display the available options. Press ENTER to accept the
desired setting, or BACK to return to the previous option
and/or exit.
Conguration Options
Ready indicator ON / OFF / or Automatic HOLD when stable
• Choose ºCelsius or ºFahrenheit
Conductivity Cell Options
Cell Constant. Choose (0.10 / 1.00 / 10.0)
Temperature Coefficient. Adjust from 0.00 - 10.0 %/ºC
Oakton Instruments
625 East Bunker Court,
Vernon Hills, IL, 60061, USA
Tel: 1-888-462-5866
Fax: 1-847-247-2984
www.4oakton.com
Oakton Instruments
625 East Bunker Court,
Vernon Hills, IL, 60061, USA
Tel: 1-888-462-5866
Fax: 1-847-247-2984
www.4oakton.com
68X651902 08/14 Rev068X651902 08/14 Rev0
BNC: pH,
mV/ORP or
Ion Selective
(PC 450 only)
ATC/Con:
6-Pin ATC /
Conductivity
or ATC
BNC: pH,
mV/ORP o
selectivo de iones
(sólo PC 450)
ATC/Con:
6-Pin ATC /
Conductividad
o ATC
A/C Power
Adapter
Output (USB or
RS232 – PC or Printer
Communication)
Transformador
A/C
Salida (USB o RS232 –
Comunicación de PC
o Impresora)
Instrument Operating Conditions
Operating Ambient Temp. 5 to 45 ºC
Operating Relative Humidity 5 to 85 %, non-condensing
Storage Temp. -20 to +60 ºC
Storage Relative Humidity 5 to 85 %, non-condensing
Pollution Degree 2
Overvoltage Category II
Weight 500 g
Size (L x W x H) 21.15 x 9.87 x 5.85 cm
Regulatory & Safety CE, TUV 3-1, FCC Class A
Power Rating DC Input: 9 VDC 1 A
Battery Requirement
2 x AA (LR6) 1.5 V batteries
(replace batteries when
battery sign blinks)
Vibration Shipping/handling per ISTA #1A
Shock
Drop test in packaging per
ISTA #1A
Enclosure (Designed To Meet) IP67 (using rubber covers)
Universal Power Adapter Operating Conditions
Operating Ambient Temp. 0 to 50 ºC
Operating Relative Humidity 0 to 90 %, non-condensing
Storage Temp. -20 to +75 ºC
Storage Relative Humidity 0 to 90 %, non-condensing
Pollution Degree 2
Overvoltage Category II
Power Rating
I/P: 100 - 240 V, 50/60 Hz, 0.3A
O/P: 9 VDC 1 A
Modelos:
Primeros pasos/Conexiones
Luego de insertar (2) baterías AA y/o conectar el suministro de energía
opcional de 110/220 VAC conecte los sensores deseados a los puertos
correspondientes.
Sondas de 12 mm y 16 mm pueden utilizar Grip-Clip™ para sujetar uno o
más sensores a un vaso y si es necesario al instrumento. El pie puede ser
extendido como se muestra arriba o puede ser montado a la pared.
Funciones del teclado
Programas de conguración
Para acceder a los ajustes abajo presione CONFIGURACIÓN, Las flechas
arriba/abajo van a hacer aparecer las opciones disponibles. Presione ENTER
para aceptar el ajuste deseado o BACK (Volver) para volver a la opción
anterior y/o salir.
Opciones de Conguración
Indicador listo ON / OFF / o HOLD (Mantenimiento) automático si
estable
• Elija ºCelsius o ºFahrenheit
Opciones de Célula de Conductividad
Constante de Célula. Elija (0.10 / 1.00 / 10.0)
Coeficiente de Temperatura. Ajuste entre 0.00 - 10.0 %/ºC
Condiciones de funcionamiento del instrumento
Temperatura de ambiente de
funcionamiento
5 a 45 ºC
Humedad relativa de
funcionamiento
5 a 85 %, no condensada
Temperatura de almacenamiento -20 a +60 ºC
Humedad relativa de
almacenamiento
5 a 85 %, no condensada
Contaminación Grado de 2
Sobretensión Categoría de II
Peso 500 g
Tamaño (L x A x A) 21,15 x 9,87 x 5,85 cm
Ordenanzas y Seguridad CE, TUV 3-1, FCC Clase A
Potencia Nominal Entrada de CC: 9 VDC 1 A
Requerimientos de batería
2 x AA (LR6) 1,5 V baterías
(reemplazar baterías cuando
parpadea el cartel de la batería)
Vibración Envío/Manejo conforme a ISTA #1A
Shock
Test de caída en envase conforme
a ISTA #1A
Carcasa (diseñado para cumplir) IP67 (utilizando cubiertas de goma)
Transformador universal Condiciones de funcionamiento
Temperatura de Ambiente de
funcionamiento
5 a 50 ºC
Humedad relativa de
funcionamiento
0 a 90 %, no condensada
Temperatura de almacenamiento -20 a +75 ºC
Humedad relativa de
almacenamiento
0 a 90 %, no condensada
Contaminación Grado de 2
Sobretensión Categoría de II
Potencia Nominal
I/P: 100 - 240 V, 50/60 Hz, 0.3A
O/P: 9 VDC 1 A
Presione una vez, para encender en el modo usado
previamente. Presione otra vez, para encender la
retroiluminación por un minuto o apagar (sólo serie 450).
Mantenga apretado por 3 segundos.
Alterne ente los modos medición y calibración. En el modo
CONFIGURACIÓN la tecla VOLVER sirve para volver a la
opción previa del menú o ajuste.
Confirmar los valores de calibración en el modo CAL.
Confirmar selecciones en el modo CONFIGURACIÓN.
Congelar o liberar el valor medido.
Personalizar los ajustes del instrumento y las preferencias.
(Vea también los programas de conguración).
Alterne entre los tipos de medición disponibles.
Guardar la medición en la memoria. Aumente el valor o
desplace hacia arriba en CONFIGURACIÓN o calibración
manual.
Volver a mostrar valores guardados en la memoria.
Reduzca el valor o desplace hacia abajo en
CONFIGURACIÓN o calibración manual.
Envíe los datos de salida a una impresora o computadora.
(Sólo serie 450).
CON 150
Con/TDS
CON 450
Con/TDS/Salinidad
PC 450
pH/mV/Con/TDS
Factor total de sólidos disueltos. Ajuste entre (0.00 - 1.00)
Temperatura de normalización. Seleccione (15.0 / 20.0 / 25.0 / 30.0 °C)
Seleccione Calibración de un punto (SPC) para aplicar un sólo valor de
calibración a todos los rangos o Calibración de varios puntos (MPC) para
calibrar rangos individuales.
Seleccione Estándares de Calibración Automáticos o Ajuste Manual
Seleccione Noticación de Calibración Pendiente
Fije la cantidad de días entre 0 y 60 para el parámetro deseado
Ver Datos de Calibración (para el parámetro que se está midiendo)
• Presione ENTER para ver cada punto que va a ser calibrado con el rango
de calibración asociado.
Ver Datos de Electrodo (para el parámetro que se está midiendo)
• Presione ENTER para ver la eficiencia del electrodo de cada punto que se
está calibrando con su rango de calibración asociado.
Ajustes del Sistema
Registro de datos:
MANUAL al apretar un botón sólo intervalo TIMED (Cronometrado). Elija
(SEG / MIN / HORA) intervalo.
Apagado automático después de 10 minutos. Elija ON oder OFF.
Ajustes del reloj:
Fecha: Elija USA (MM/DD/AAAA) o Euro (DD/MM/AAAA).
Tiempo: Elija (24HR o 12HR). Si 12HR elija AM o PM.
Ajustar tipo de impresora:
CSV (Valores Separados por Coma) – mejor formato para la computadora
Impresora (texto) – mejor formato para impresora.
Elija Manual (MAN) al apretar un botón o intervalo TIMED
(Cronometrado).
Si cronometrado, elija (SEG / MIN / HORA).
Resetear
NO (NRO). de salidas de las opciones del menú de reseteo sin acción.
FACTORY RESET (RESETEO DE FÁBRICA). Retorna todos los ajustes von
excepción de fecha/tiempo y calibración ATC a los valores estándares de
fábrica luego de oprimir ENTER el medidor es reseteado.
DATA RESET (RESETEO DE DATOS). Elimina datos guardados en la
memoria mientras otros ajustes se conservan luego de oprimir ENTER.
CALIBRATION RESET (RESETEO DE CALIBRACIÓN). Elimina datos no
pertenecientes a la calibración ATC mientras otros ajustes se conservan
luego de oprimir ENTER.
Conductividad/TDS/Salinidad Calibración
Para mejores resultados se recomienda una calibración periódica con
estándares conocidos y precisos. Calibre con estándares cercanos a su
rango de medición deseado. Revuelva para los mejores resultados. Luego
de la calibración va a ser visible en la visualización inferior la eficiencia del
electrodo que corresponde a la medición activa. “- - -“ va a aparecer si no
se lleva a cabo ninguna calibración. Presione MEAS para regresar al modo
de medición.
Utilizando Reconocimiento Automático de Conductividad
1. Desde el modo de medición de conductividad sumerja el sensor en
su estándar – 84 μS, 1413 μS, 12,88 mS, o 111,8 mS, luego presione
CAL. La visualización primaria va a buscar al estándar más cercano.
La visualización secundaria va a mostrar el valor no ajustado y la
visualización inferior va a mostrar la temperatura. Se permite un valor de
calibración en cada rango.
2. Cuando aparece el indicador LISTO presione ENTER para aceptar. En la
medición primaria va a brillar LISTO”.
3. Para calibrar otro estándar, lave sus electrodos y luego sumérjalos en su
próximo estándar o presione MEAS para regresar al modo de medición.
La pantalla primaria va a buscar al valor estándar más cercano, que no
ha sido calibrado aún, mientras que la pantalla secundaria va a mostrar
el valor no ajustado. Cuando aparece el indicador LISTO presione
ENTER para aceptar.
4. Luego de que se haya calibrado la cantidad deseada de estándares, la
visualización inferior va a indicar la eficiencia en el modo de medición.
Utilizando Ajuste Manual
1. Desde el modo de medición Conductividad/TDS/SALINIDAD sumerja el
sensor en su estándar y luego presione CAL.
2. Cuando aparezca el indicador LISTO utilize las flechas arriba/abajo para
ajustar la medición primaria para que sea igual que el valor estándar a la
temperatura medida, luego presione ENTER.
Rango
#
Rango de
Conductividad
Valores de Calibración Automáticos
Temperatura de Normalización
25 ºC 20 ºC
r 1 0.00 - 19.99 µS Ninguno Ninguno
r 2 20.0 - 199.9 µS 84 µS 76 µS
r 3 200.0 - 1999 µS 1413 µS 1278 µS
r 4 2.00 - 19.99 mS 12.88 mS 11.67 mS
r 5 20.0 - 200.0 mS 111.8 mS 102.1 mS
3. Lave sus electrodos y sumerja el sensor en su próximo estándar o
presione MEAS para regresar al modo de medición, Repita si es
necesario. RLuego de que se haya calibrado la cantidad deseada de
estándares, la visualización inferior va a indicar la eficiencia en el modo
de medición.
Calibración de temperatura/ATC Manual
La calibración de la temperatura se recomiendo antes del primer uso, luego
del reemplazo del sensor ATC y periódicamente según sea necesario.
1. Oprima CAL de cualquier medición, luego oprima MODE (Modo).
2. Siga con el paso 3 por una ATC manual, de otra manera sumerja el sensor
de temperatura en una solución con una temperatura determinada
conocida. La pantalla superior muestra la temperatura activa mientras
que la pantalla inferior muestra la temperatura estándar de fábrica sin
ajuste.
3. Use las flechas arriba/abajo para ajustar el monitor superior.
Oprima ENTER para aceptar la temperatura de calibración.
El valor ajustable máximo es ±10 ºC (o ±18 ºF) desde el valor de fábrica.
Mensajes de error
“ERR” va a aparecer si existe una condición de error o si es presionada
la tecla equivocada. Ejemplos comunes son entre otros:
• Apretar ENTER durante la calibración antes de que aparezca la
indicación “READY (Listo). Espere a que aparezca la indicación READY
(Listo) antes de apretar ENTER.
RI (Rango inferior) • RS (Rango superior)
Uso intencionado, mantenimiento y precauciones
Estos medidores portables utilizan sensores para detectar varios
parámetros para mediciones basadas en agua. Para el mantenimiento de
rutina desconecte el cable de alimentación o la batería y luego pase un
trapo húmedo por la pantalla. Si es necesario puede utilizarse agua tibia o
un detergente suave basado en agua. Remueve inmediatamente cualquier
substancia derramada sobre el medidor con el procedimiento de limpieza
correcto para el tipo de substancia derramada.
No utilice este equipo en atmósferas potencialmente peligrosas.
Vea las instrucciones del electrodo para su uso, almacenamiento y
limpieza.
Asegúrese de que ningún liquido ingrese al instrumento.
• No use químicos de limpieza agresivos (solventes o substancias
similares).
No hay partes en el interior que pueden ser mantenidos por el usuario.
El intento de mantener partes internas puede anular la garantía.
ADVERTENCIA: No está permitido ninguna modificación de este equipo.
2. When the READY indicator appears use up/down arrows to
adjust the primary reading to match the standard value at
the measured temperature, then press ENTER.
3. Rinse your electrode(s) then dip the sensor into your next
standard, or press MEAS to return to measurement mode.
Repeat as necessary. After the desired number of standards
have been calibrated, the bottom display will show the
efficiency in measurement mode.
Temperature Calibration/Manual ATC
Temperature calibration is recommended prior to first use,
after ATC sensor replacement, and periodically as needed.
1. Press CAL from any measurement, then press MODE.
2. Skip to step 3 for manual ATC, otherwise, dip the temperature
sensor into a solution with a known accurate temperature.
The upper display shows the active temperature while
the lower display shows the factory default temperature
without adjustment.
3. Use up/down arrows to adjust the upper display.
Press ENTER to accept the calibration temperature.
The maximum adjustable value is ±10 ºC (or ±18 ºF) from
factory default.
Error Messages
“ERR” will appear when an error condition exists or the
incorrect key is pressed. Common examples include:
• Pressing ENTER during calibration before the READY
indicator appears. Wait for the READY indicator before
pressing ENTER.
UR (Under Range) • OR (Over Range)
Intended Use, Maintenance & Precautions
These handheld meters use sensors to detect various parameters
for water-based measurements. For routine maintenance
disconnect the power cord or battery, then dust or wipe the
display using a damp cloth. If necessary, warm water or a mild
water based detergent can be used. Immediately remove any
spilled substance from contact with the meter using the proper
cleaning procedure for the type of spill.
Do not use this equipment in potentially explosive atmospheres.
Refer to the electrode instructions for use, storage and cleaning.
Ensure that no liquid enters the instrument.
Do not use any aggressive cleaning chemicals (solvents or
similar agents).
• There are no user serviceable parts inside. Attempts to
service internal parts may void the warranty.
WARNING: No modification of this equipment is allowed.
Total Dissolved Solids Factor. Adjust from (0.00 - 1.00)
Normalization Temperature. Choose (15.0 / 20.0 / 25.0 /
30.0 °C)
• Choose Single-Point Calibration (SPC) to apply a single
calibration value across all ranges or Multi-Point Calibration
(MPC) to calibrate individual range(s).
Choose Automatic Calibration Standards or Manual Adjustment
Select Calibration Due Reminder
Set number of days from 0-60 for desired parameter
View Calibration Data (for the parameter being measured)
• Press ENTER to view each point that is calibrated with the
associated calibration range.
View Electrode Data (for the parameter being measured)
• Press ENTER to view the electrode efficiency of each point
that is calibrated with the associated calibration range.
System Settings
Data Logging:
MANUAL upon key press only
TIMED interval. Choose (SEC / MIN / HOUR) interval.
Automatic shut off after 10 minutes. Choose ON or OFF.
Clock Settings:
Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY).
Time: Choose (24HR or 12HR). If 12HR, choose AM or PM.
Set Printer Type:
CSV (Comma Separated Values) – best format for computer
Printer (Text) – best format for printer.
Choose Manual (MAN) upon key press or TIMED interval.
If timed, choose (SEC / MIN / HOUR).
Reset
NO. Exits from reset menu options without action.
FACTORY RESET. Returns all settings except date/time
and ATC calibration to factory default values after ENTER is
pressed then restarts meter.
DATA RESET. Erases data stored in memory while retaining
other settings after ENTER is pressed.
CALIBRATION RESET. Erases non-ATC calibration data while
retaining other settings after ENTER is pressed.
Conductivity/TDS/Salinity Calibration
For best results, periodic calibration with known accurate
standards is recommended. Calibrate with standard(s) near
your intended measuring range. Provide stirring for best results.
After calibration, the electrode efficiency that corresponds to
the active measurement will be visible on the bottom display.
“- - -“ will be shown if no calibration is performed. Press MEAS
to return to measurement mode at any time.
Using Automatic Conductivity Recognition
1. From Conductivity measurement mode, dip the sensor
into your standard – either 84 μS, 1413 μS, 12.88 mS,
or 111.8 mS, then press CAL. The primary display will search
for the nearest standard. The secondary display will show
the un-adjusted value and the bottom display will show the
temperature. One calibration value in each range is allowed.
2. When the READY indicator appears, press ENTER to accept.
The primary reading will flash DONE”.
3. To calibrate another standard, rinse your electrode(s) then
dip into your the next standard or press MEAS to return
to measurement mode at any time. The primary display
will search for the nearest standard value that has not yet
been calibrated, while the secondary display will show
the unadjusted value. When the READY indicator appears,
press ENTER to accept.
4. After the desired number of standards has been calibrated,
the bottom display will show the efficiency in measurement
mode.
Using Manual Adjustment
1. From Conductivity/TDS/SALINITY measurement mode, dip
the sensor into your standard then press CAL.
150 & 450 Series Waterproof Handheld Meter
Conductivity/TDS/Salinity Operation Instructions
150 & 450 Series Medidor de mano resistente al agua
Conductividad/TDS/Salinidad Instrucciones de funcionamiento
Part of Thermo Fisher Scientific Parte de Thermo Fisher Scientific
CON 150
Con/TDS
CON 450
Con/TDS/Salinity
PC 450
pH/mV/Con/TDS
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-16/24, Singapore 139949
Tel: (65) 6778-6876
Fax: (65) 6778-0086
eutech@thermofisher.com
www.eutechinst.com
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-16/24, Singapore 139949
Tel: (65) 6778-6876
Fax: (65) 6778-0086
eutech@thermofisher.com
www.eutechinst.com
Press once to power ON in the mode that was
previously used. Press again to turn backlight on
for one minute or off (450 series only). Hold for 3
seconds to power OFF.
Toggle between measurement and calibration
modes. In SETUP mode, BACK serves to return to
the previous menu option or setting.
Confirm calibration values in CAL mode.
Confirm selections in SETUP mode.
Freeze or release the measured reading.
Customize instrument settings and preferences.
(See also Setup Programs)
Toggle between available measurement types.
Save measurement into memory. Increase value
or scroll up in SETUP or manual calibration.
Recall saved values from memory. Decrease value
or scroll down in SETUP or manual calibration.
Send output data to printer or computer.
(450 series only).
Range
#
Conductivity
Range
Automatic Calibration Values
Normalization Temperature
25 ºC 20 ºC
r 1 0.00 - 19.99 µS None None
r 2 20.0 - 199.9 µS 84 µS 76 µS
r 3 200.0 - 1999 µS 1413 µS 1278 µS
r 4 2.00 - 19.99 mS 12.88 mS 11.67 mS
r 5 20.0 - 200.0 mS 111.8 mS 102.1 mS
Models:
Getting Started/Connections
After installing (2) AA batteries and/or connecting the optional
110/220 VAC power supply, connect the desired sensors to the
corresponding ports.
12 mm and 16 mm probes can utilize the Grip-Clip™ to attach
one or more sensors to a beaker and to the instrument as
needed. The stand can be extended as shown above or used
for wall-mounting.
Keypad Functions
Setup Programs
To access the settings below, press SETUP. Up/down arrows
will display the available options. Press ENTER to accept the
desired setting, or BACK to return to the previous option
and/or exit.
Conguration Options
Ready indicator ON / OFF / or Automatic HOLD when stable
• Choose ºCelsius or ºFahrenheit
Conductivity Cell Options
Cell Constant. Choose (0.10 / 1.00 / 10.0)
Temperature Coefficient. Adjust from 0.00 - 10.0 %/ºC
Oakton Instruments
625 East Bunker Court,
Vernon Hills, IL, 60061, USA
Tel: 1-888-462-5866
Fax: 1-847-247-2984
www.4oakton.com
Oakton Instruments
625 East Bunker Court,
Vernon Hills, IL, 60061, USA
Tel: 1-888-462-5866
Fax: 1-847-247-2984
www.4oakton.com
68X651902 08/14 Rev068X651902 08/14 Rev0
BNC: pH,
mV/ORP or
Ion Selective
(PC 450 only)
ATC/Con:
6-Pin ATC /
Conductivity
or ATC
BNC: pH,
mV/ORP o
selectivo de iones
(sólo PC 450)
ATC/Con:
6-Pin ATC /
Conductividad
o ATC
A/C Power
Adapter
Output (USB or
RS232 – PC or Printer
Communication)
Transformador
A/C
Salida (USB o RS232 –
Comunicación de PC
o Impresora)
Instrument Operating Conditions
Operating Ambient Temp. 5 to 45 ºC
Operating Relative Humidity 5 to 85 %, non-condensing
Storage Temp. -20 to +60 ºC
Storage Relative Humidity 5 to 85 %, non-condensing
Pollution Degree 2
Overvoltage Category II
Weight 500 g
Size (L x W x H) 21.15 x 9.87 x 5.85 cm
Regulatory & Safety CE, TUV 3-1, FCC Class A
Power Rating DC Input: 9 VDC 1 A
Battery Requirement
2 x AA (LR6) 1.5 V batteries
(replace batteries when
battery sign blinks)
Vibration Shipping/handling per ISTA #1A
Shock
Drop test in packaging per
ISTA #1A
Enclosure (Designed To Meet) IP67 (using rubber covers)
Universal Power Adapter Operating Conditions
Operating Ambient Temp. 0 to 50 ºC
Operating Relative Humidity 0 to 90 %, non-condensing
Storage Temp. -20 to +75 ºC
Storage Relative Humidity 0 to 90 %, non-condensing
Pollution Degree 2
Overvoltage Category II
Power Rating
I/P: 100 - 240 V, 50/60 Hz, 0.3A
O/P: 9 VDC 1 A
Modelos:
Primeros pasos/Conexiones
Luego de insertar (2) baterías AA y/o conectar el suministro de energía
opcional de 110/220 VAC conecte los sensores deseados a los puertos
correspondientes.
Sondas de 12 mm y 16 mm pueden utilizar Grip-Clip™ para sujetar uno o
más sensores a un vaso y si es necesario al instrumento. El pie puede ser
extendido como se muestra arriba o puede ser montado a la pared.
Funciones del teclado
Programas de conguración
Para acceder a los ajustes abajo presione CONFIGURACIÓN, Las flechas
arriba/abajo van a hacer aparecer las opciones disponibles. Presione ENTER
para aceptar el ajuste deseado o BACK (Volver) para volver a la opción
anterior y/o salir.
Opciones de Conguración
Indicador listo ON / OFF / o HOLD (Mantenimiento) automático si
estable
• Elija ºCelsius o ºFahrenheit
Opciones de Célula de Conductividad
Constante de Célula. Elija (0.10 / 1.00 / 10.0)
Coeficiente de Temperatura. Ajuste entre 0.00 - 10.0 %/ºC
Condiciones de funcionamiento del instrumento
Temperatura de ambiente de
funcionamiento
5 a 45 ºC
Humedad relativa de
funcionamiento
5 a 85 %, no condensada
Temperatura de almacenamiento -20 a +60 ºC
Humedad relativa de
almacenamiento
5 a 85 %, no condensada
Contaminación Grado de 2
Sobretensión Categoría de II
Peso 500 g
Tamaño (L x A x A) 21,15 x 9,87 x 5,85 cm
Ordenanzas y Seguridad CE, TUV 3-1, FCC Clase A
Potencia Nominal Entrada de CC: 9 VDC 1 A
Requerimientos de batería
2 x AA (LR6) 1,5 V baterías
(reemplazar baterías cuando
parpadea el cartel de la batería)
Vibración Envío/Manejo conforme a ISTA #1A
Shock
Test de caída en envase conforme
a ISTA #1A
Carcasa (diseñado para cumplir) IP67 (utilizando cubiertas de goma)
Transformador universal Condiciones de funcionamiento
Temperatura de Ambiente de
funcionamiento
5 a 50 ºC
Humedad relativa de
funcionamiento
0 a 90 %, no condensada
Temperatura de almacenamiento -20 a +75 ºC
Humedad relativa de
almacenamiento
0 a 90 %, no condensada
Contaminación Grado de 2
Sobretensión Categoría de II
Potencia Nominal
I/P: 100 - 240 V, 50/60 Hz, 0.3A
O/P: 9 VDC 1 A
Presione una vez, para encender en el modo usado
previamente. Presione otra vez, para encender la
retroiluminación por un minuto o apagar (sólo serie 450).
Mantenga apretado por 3 segundos.
Alterne ente los modos medición y calibración. En el modo
CONFIGURACIÓN la tecla VOLVER sirve para volver a la
opción previa del menú o ajuste.
Confirmar los valores de calibración en el modo CAL.
Confirmar selecciones en el modo CONFIGURACIÓN.
Congelar o liberar el valor medido.
Personalizar los ajustes del instrumento y las preferencias.
(Vea también los programas de conguración).
Alterne entre los tipos de medición disponibles.
Guardar la medición en la memoria. Aumente el valor o
desplace hacia arriba en CONFIGURACIÓN o calibración
manual.
Volver a mostrar valores guardados en la memoria.
Reduzca el valor o desplace hacia abajo en
CONFIGURACIÓN o calibración manual.
Envíe los datos de salida a una impresora o computadora.
(Sólo serie 450).
CON 150
Con/TDS
CON 450
Con/TDS/Salinidad
PC 450
pH/mV/Con/TDS
Factor total de sólidos disueltos. Ajuste entre (0.00 - 1.00)
Temperatura de normalización. Seleccione (15.0 / 20.0 / 25.0 / 30.0 °C)
Seleccione Calibración de un punto (SPC) para aplicar un sólo valor de
calibración a todos los rangos o Calibración de varios puntos (MPC) para
calibrar rangos individuales.
Seleccione Estándares de Calibración Automáticos o Ajuste Manual
Seleccione Noticación de Calibración Pendiente
Fije la cantidad de días entre 0 y 60 para el parámetro deseado
Ver Datos de Calibración (para el parámetro que se está midiendo)
• Presione ENTER para ver cada punto que va a ser calibrado con el rango
de calibración asociado.
Ver Datos de Electrodo (para el parámetro que se está midiendo)
• Presione ENTER para ver la eficiencia del electrodo de cada punto que se
está calibrando con su rango de calibración asociado.
Ajustes del Sistema
Registro de datos:
MANUAL al apretar un botón sólo intervalo TIMED (Cronometrado). Elija
(SEG / MIN / HORA) intervalo.
Apagado automático después de 10 minutos. Elija ON oder OFF.
Ajustes del reloj:
Fecha: Elija USA (MM/DD/AAAA) o Euro (DD/MM/AAAA).
Tiempo: Elija (24HR o 12HR). Si 12HR elija AM o PM.
Ajustar tipo de impresora:
CSV (Valores Separados por Coma) – mejor formato para la computadora
Impresora (texto) – mejor formato para impresora.
Elija Manual (MAN) al apretar un botón o intervalo TIMED
(Cronometrado).
Si cronometrado, elija (SEG / MIN / HORA).
Resetear
NO (NRO). de salidas de las opciones del menú de reseteo sin acción.
FACTORY RESET (RESETEO DE FÁBRICA). Retorna todos los ajustes von
excepción de fecha/tiempo y calibración ATC a los valores estándares de
fábrica luego de oprimir ENTER el medidor es reseteado.
DATA RESET (RESETEO DE DATOS). Elimina datos guardados en la
memoria mientras otros ajustes se conservan luego de oprimir ENTER.
CALIBRATION RESET (RESETEO DE CALIBRACIÓN). Elimina datos no
pertenecientes a la calibración ATC mientras otros ajustes se conservan
luego de oprimir ENTER.
Conductividad/TDS/Salinidad Calibración
Para mejores resultados se recomienda una calibración periódica con
estándares conocidos y precisos. Calibre con estándares cercanos a su
rango de medición deseado. Revuelva para los mejores resultados. Luego
de la calibración va a ser visible en la visualización inferior la eficiencia del
electrodo que corresponde a la medición activa. “- - -“ va a aparecer si no
se lleva a cabo ninguna calibración. Presione MEAS para regresar al modo
de medición.
Utilizando Reconocimiento Automático de Conductividad
1. Desde el modo de medición de conductividad sumerja el sensor en
su estándar – 84 μS, 1413 μS, 12,88 mS, o 111,8 mS, luego presione
CAL. La visualización primaria va a buscar al estándar más cercano.
La visualización secundaria va a mostrar el valor no ajustado y la
visualización inferior va a mostrar la temperatura. Se permite un valor de
calibración en cada rango.
2. Cuando aparece el indicador LISTO presione ENTER para aceptar. En la
medición primaria va a brillar LISTO”.
3. Para calibrar otro estándar, lave sus electrodos y luego sumérjalos en su
próximo estándar o presione MEAS para regresar al modo de medición.
La pantalla primaria va a buscar al valor estándar más cercano, que no
ha sido calibrado aún, mientras que la pantalla secundaria va a mostrar
el valor no ajustado. Cuando aparece el indicador LISTO presione
ENTER para aceptar.
4. Luego de que se haya calibrado la cantidad deseada de estándares, la
visualización inferior va a indicar la eficiencia en el modo de medición.
Utilizando Ajuste Manual
1. Desde el modo de medición Conductividad/TDS/SALINIDAD sumerja el
sensor en su estándar y luego presione CAL.
2. Cuando aparezca el indicador LISTO utilize las flechas arriba/abajo para
ajustar la medición primaria para que sea igual que el valor estándar a la
temperatura medida, luego presione ENTER.
Rango
#
Rango de
Conductividad
Valores de Calibración Automáticos
Temperatura de Normalización
25 ºC 20 ºC
r 1 0.00 - 19.99 µS Ninguno Ninguno
r 2 20.0 - 199.9 µS 84 µS 76 µS
r 3 200.0 - 1999 µS 1413 µS 1278 µS
r 4 2.00 - 19.99 mS 12.88 mS 11.67 mS
r 5 20.0 - 200.0 mS 111.8 mS 102.1 mS
3. Lave sus electrodos y sumerja el sensor en su próximo estándar o
presione MEAS para regresar al modo de medición, Repita si es
necesario. RLuego de que se haya calibrado la cantidad deseada de
estándares, la visualización inferior va a indicar la eficiencia en el modo
de medición.
Calibración de temperatura/ATC Manual
La calibración de la temperatura se recomiendo antes del primer uso, luego
del reemplazo del sensor ATC y periódicamente según sea necesario.
1. Oprima CAL de cualquier medición, luego oprima MODE (Modo).
2. Siga con el paso 3 por una ATC manual, de otra manera sumerja el sensor
de temperatura en una solución con una temperatura determinada
conocida. La pantalla superior muestra la temperatura activa mientras
que la pantalla inferior muestra la temperatura estándar de fábrica sin
ajuste.
3. Use las flechas arriba/abajo para ajustar el monitor superior.
Oprima ENTER para aceptar la temperatura de calibración.
El valor ajustable máximo es ±10 ºC (o ±18 ºF) desde el valor de fábrica.
Mensajes de error
“ERR” va a aparecer si existe una condición de error o si es presionada
la tecla equivocada. Ejemplos comunes son entre otros:
• Apretar ENTER durante la calibración antes de que aparezca la
indicación “READY (Listo). Espere a que aparezca la indicación READY
(Listo) antes de apretar ENTER.
RI (Rango inferior) • RS (Rango superior)
Uso intencionado, mantenimiento y precauciones
Estos medidores portables utilizan sensores para detectar varios
parámetros para mediciones basadas en agua. Para el mantenimiento de
rutina desconecte el cable de alimentación o la batería y luego pase un
trapo húmedo por la pantalla. Si es necesario puede utilizarse agua tibia o
un detergente suave basado en agua. Remueve inmediatamente cualquier
substancia derramada sobre el medidor con el procedimiento de limpieza
correcto para el tipo de substancia derramada.
No utilice este equipo en atmósferas potencialmente peligrosas.
Vea las instrucciones del electrodo para su uso, almacenamiento y
limpieza.
Asegúrese de que ningún liquido ingrese al instrumento.
• No use químicos de limpieza agresivos (solventes o substancias
similares).
No hay partes en el interior que pueden ser mantenidos por el usuario.
El intento de mantener partes internas puede anular la garantía.
ADVERTENCIA: No está permitido ninguna modificación de este equipo.

Transcripción de documentos

Part of Thermo Fisher Scientific 150 & 450 Series Waterproof Handheld Meter 150 & 450 Series Medidor de mano resistente al agua Conductivity/TDS/Salinity Operation Instructions Models: CON 150 Con/TDS CON 450 Con/TDS/Salinity PC 450 pH/mV/Con/TDS Getting Started/Connections After installing (2) AA batteries and/or connecting the optional 110/220 VAC power supply, connect the desired sensors to the corresponding ports. BNC: pH, mV/ORP or Ion Selective (PC 450 only) ATC/Con: 6-Pin ATC / Conductivity or ATC A/C Power Adapter Output (USB or RS232 – PC or Printer Communication) 12 mm and 16 mm probes can utilize the Grip-Clip™ to attach one or more sensors to a beaker and to the instrument as needed. The stand can be extended as shown above or used for wall-mounting. Keypad Functions Press once to power ON in the mode that was previously used. Press again to turn backlight on for one minute or off (450 series only). Hold for 3 seconds to power OFF. Toggle between measurement and calibration modes. In SETUP mode, BACK serves to return to the previous menu option or setting. Confirm calibration values in CAL mode. Confirm selections in SETUP mode. Freeze or release the measured reading. Customize instrument settings and preferences. (See also Setup Programs) Toggle between available measurement types. Save measurement into memory. Increase value or scroll up in SETUP or manual calibration. Recall saved values from memory. Decrease value or scroll down in SETUP or manual calibration. Send output data to printer or computer. (450 series only). Setup Programs To access the settings below, press SETUP. Up/down arrows will display the available options. Press ENTER to accept the desired setting, or BACK to return to the previous option and/or exit. Configuration Options • Ready indicator ON / OFF / or Automatic HOLD when stable • Choose ºCelsius or ºFahrenheit Conductivity Cell Options • Cell Constant. Choose (0.10 / 1.00 / 10.0) • Temperature Coefficient. Adjust from 0.00 - 10.0 %/ºC • Total Dissolved Solids Factor. Adjust from (0.00 - 1.00) • Normalization Temperature. Choose (15.0 / 20.0 / 25.0 / 30.0 °C) • Choose Single-Point Calibration (SPC) to apply a single calibration value across all ranges or Multi-Point Calibration (MPC) to calibrate individual range(s). • Choose Automatic Calibration Standards or Manual Adjustment Select Calibration Due Reminder • Set number of days from 0-60 for desired parameter View Calibration Data (for the parameter being measured) • Press ENTER to view each point that is calibrated with the associated calibration range. View Electrode Data (for the parameter being measured) • Press ENTER to view the electrode efficiency of each point that is calibrated with the associated calibration range. System Settings • Data Logging: MANUAL upon key press only TIMED interval. Choose (SEC / MIN / HOUR) interval. • Automatic shut off after 10 minutes. Choose ON or OFF. • Clock Settings: Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY). Time: Choose (24HR or 12HR). If 12HR, choose AM or PM. • Set Printer Type: CSV (Comma Separated Values) – best format for computer Printer (Text) – best format for printer. Choose Manual (MAN) upon key press or TIMED interval. If timed, choose (SEC / MIN / HOUR). Reset • NO. Exits from reset menu options without action. • FACTORY RESET. Returns all settings except date/time and ATC calibration to factory default values after ENTER is pressed then restarts meter. • DATA RESET. Erases data stored in memory while retaining other settings after ENTER is pressed. • CALIBRATION RESET. Erases non-ATC calibration data while retaining other settings after ENTER is pressed. Conductivity/TDS/Salinity Calibration For best results, periodic calibration with known accurate standards is recommended. Calibrate with standard(s) near your intended measuring range. Provide stirring for best results. After calibration, the electrode efficiency that corresponds to the active measurement will be visible on the bottom display. “- - -“ will be shown if no calibration is performed. Press MEAS to return to measurement mode at any time. Range # Conductivity Range r1 r2 r3 r4 r5 0.00 - 19.99 µS 20.0 - 199.9 µS 200.0 - 1999 µS 2.00 - 19.99 mS 20.0 - 200.0 mS Automatic Calibration Values Normalization Temperature 25 ºC 20 ºC None None 84 µS 76 µS 1413 µS 1278 µS 12.88 mS 11.67 mS 111.8 mS 102.1 mS Using Automatic Conductivity Recognition 1. From Conductivity measurement mode, dip the sensor into your standard – either 84 μS, 1413 μS, 12.88 mS, or 111.8 mS, then press CAL. The primary display will search for the nearest standard. The secondary display will show the un-adjusted value and the bottom display will show the temperature. One calibration value in each range is allowed. 2. When the READY indicator appears, press ENTER to accept. The primary reading will flash “DONE”. 3. To calibrate another standard, rinse your electrode(s) then dip into your the next standard or press MEAS to return to measurement mode at any time. The primary display will search for the nearest standard value that has not yet been calibrated, while the secondary display will show the unadjusted value. When the READY indicator appears, press ENTER to accept. 4. After the desired number of standards has been calibrated, the bottom display will show the efficiency in measurement mode. Using Manual Adjustment 1. From Conductivity/TDS/SALINITY measurement mode, dip the sensor into your standard then press CAL. Parte de Thermo Fisher Scientific Conductividad/TDS/Salinidad Instrucciones de funcionamiento 2. When the READY indicator appears use up/down arrows to adjust the primary reading to match the standard value at the measured temperature, then press ENTER. 3. Rinse your electrode(s) then dip the sensor into your next standard, or press MEAS to return to measurement mode. Repeat as necessary. After the desired number of standards have been calibrated, the bottom display will show the efficiency in measurement mode. Modelos: Temperature Calibration/Manual ATC Temperature calibration is recommended prior to first use, after ATC sensor replacement, and periodically as needed. 1. Press CAL from any measurement, then press MODE. 2. Skip to step 3 for manual ATC, otherwise, dip the temperature sensor into a solution with a known accurate temperature. The upper display shows the active temperature while the lower display shows the factory default temperature without adjustment. 3. Use up/down arrows to adjust the upper display. Press ENTER to accept the calibration temperature. The maximum adjustable value is ±10 ºC (or ±18 ºF) from factory default. Error Messages CON 150 Con/TDS CON 450 Con/TDS/Salinidad Primeros pasos/Conexiones Luego de insertar (2) baterías AA y/o conectar el suministro de energía opcional de 110/220 VAC conecte los sensores deseados a los puertos correspondientes. BNC: pH, mV/ORP o selectivo de iones (sólo PC 450) “ERR” will appear when an error condition exists or the incorrect key is pressed. Common examples include: • Pressing ENTER during calibration before the “READY” indicator appears. Wait for the “READY” indicator before pressing ENTER. • UR (Under Range) • OR (Over Range) Intended Use, Maintenance & Precautions These handheld meters use sensors to detect various parameters for water-based measurements. For routine maintenance disconnect the power cord or battery, then dust or wipe the display using a damp cloth. If necessary, warm water or a mild water based detergent can be used. Immediately remove any spilled substance from contact with the meter using the proper cleaning procedure for the type of spill. • Do not use this equipment in potentially explosive atmospheres. • Refer to the electrode instructions for use, storage and cleaning. • Ensure that no liquid enters the instrument. • Do not use any aggressive cleaning chemicals (solvents or similar agents). • There are no user serviceable parts inside. Attempts to service internal parts may void the warranty. • WARNING: No modification of this equipment is allowed. Instrument Operating Conditions Operating Ambient Temp. 5 to 45 ºC Operating Relative Humidity 5 to 85 %, non-condensing Storage Temp. -20 to +60 ºC Storage Relative Humidity 5 to 85 %, non-condensing Pollution Degree 2 Overvoltage Category II Weight 500 g Size (L x W x H) 21.15 x 9.87 x 5.85 cm Regulatory & Safety CE, TUV 3-1, FCC Class A Power Rating DC Input: 9 VDC 1 A 2 x AA (LR6) 1.5 V batteries Battery Requirement (replace batteries when battery sign blinks) Vibration Shipping/handling per ISTA #1A Drop test in packaging per Shock ISTA #1A Enclosure (Designed To Meet) IP67 (using rubber covers) Universal Power Adapter Operating Conditions Operating Ambient Temp. 0 to 50 ºC Operating Relative Humidity 0 to 90 %, non-condensing Storage Temp. -20 to +75 ºC Storage Relative Humidity 0 to 90 %, non-condensing Pollution Degree 2 Overvoltage Category II I/P: 100 - 240 V, 50/60 Hz, 0.3A Power Rating O/P: 9 VDC 1 A Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24, Singapore 139949 Tel: (65) 6778-6876 Fax: (65) 6778-0086 [email protected] www.eutechinst.com Oakton Instruments 625 East Bunker Court, Vernon Hills, IL, 60061, USA Tel: 1-888-462-5866 Fax: 1-847-247-2984 [email protected] www.4oakton.com 68X651902 08/14 Rev0 PC 450 pH/mV/Con/TDS ATC/Con: 6-Pin ATC / Conductividad o ATC Transformador A/C Salida (USB o RS232 – Comunicación de PC o Impresora) Sondas de 12 mm y 16 mm pueden utilizar Grip-Clip™ para sujetar uno o más sensores a un vaso y si es necesario al instrumento. El pie puede ser extendido como se muestra arriba o puede ser montado a la pared. Funciones del teclado Presione una vez, para encender en el modo usado previamente. Presione otra vez, para encender la retroiluminación por un minuto o apagar (sólo serie 450). Mantenga apretado por 3 segundos. Alterne ente los modos medición y calibración. En el modo CONFIGURACIÓN la tecla VOLVER sirve para volver a la opción previa del menú o ajuste. Confirmar los valores de calibración en el modo CAL. Confirmar selecciones en el modo CONFIGURACIÓN. Congelar o liberar el valor medido. Personalizar los ajustes del instrumento y las preferencias. (Vea también los programas de configuración). Alterne entre los tipos de medición disponibles. Guardar la medición en la memoria. Aumente el valor o desplace hacia arriba en CONFIGURACIÓN o calibración manual. Volver a mostrar valores guardados en la memoria. Reduzca el valor o desplace hacia abajo en CONFIGURACIÓN o calibración manual. Envíe los datos de salida a una impresora o computadora. (Sólo serie 450). Programas de configuración Para acceder a los ajustes abajo presione CONFIGURACIÓN, Las flechas arriba/abajo van a hacer aparecer las opciones disponibles. Presione ENTER para aceptar el ajuste deseado o BACK (Volver) para volver a la opción anterior y/o salir. Opciones de Configuración • Indicador listo ON / OFF / o HOLD (Mantenimiento) automático si estable • Elija ºCelsius o ºFahrenheit Opciones de Célula de Conductividad • Constante de Célula. Elija (0.10 / 1.00 / 10.0) • Coeficiente de Temperatura. Ajuste entre 0.00 - 10.0 %/ºC • Factor total de sólidos disueltos. Ajuste entre (0.00 - 1.00) • Temperatura de normalización. Seleccione (15.0 / 20.0 / 25.0 / 30.0 °C) • Seleccione Calibración de un punto (SPC) para aplicar un sólo valor de calibración a todos los rangos o Calibración de varios puntos (MPC) para calibrar rangos individuales. • Seleccione Estándares de Calibración Automáticos o Ajuste Manual Seleccione Notificación de Calibración Pendiente • Fije la cantidad de días entre 0 y 60 para el parámetro deseado Ver Datos de Calibración (para el parámetro que se está midiendo) • Presione ENTER para ver cada punto que va a ser calibrado con el rango de calibración asociado. Ver Datos de Electrodo (para el parámetro que se está midiendo) • Presione ENTER para ver la eficiencia del electrodo de cada punto que se está calibrando con su rango de calibración asociado. Ajustes del Sistema • Registro de datos: MANUAL al apretar un botón sólo intervalo TIMED (Cronometrado). Elija (SEG / MIN / HORA) intervalo. • Apagado automático después de 10 minutos. Elija ON oder OFF. • Ajustes del reloj: Fecha: Elija USA (MM/DD/AAAA) o Euro (DD/MM/AAAA). Tiempo: Elija (24HR o 12HR). Si 12HR elija AM o PM. • Ajustar tipo de impresora: CSV (Valores Separados por Coma) – mejor formato para la computadora Impresora (texto) – mejor formato para impresora. Elija Manual (MAN) al apretar un botón o intervalo TIMED (Cronometrado). Si cronometrado, elija (SEG / MIN / HORA). Resetear • NO (NRO). de salidas de las opciones del menú de reseteo sin acción. • FACTORY RESET (RESETEO DE FÁBRICA). Retorna todos los ajustes von excepción de fecha/tiempo y calibración ATC a los valores estándares de fábrica luego de oprimir ENTER el medidor es reseteado. • DATA RESET (RESETEO DE DATOS). Elimina datos guardados en la memoria mientras otros ajustes se conservan luego de oprimir ENTER. • CALIBRATION RESET (RESETEO DE CALIBRACIÓN). Elimina datos no pertenecientes a la calibración ATC mientras otros ajustes se conservan luego de oprimir ENTER. Conductividad/TDS/Salinidad Calibración Para mejores resultados se recomienda una calibración periódica con estándares conocidos y precisos. Calibre con estándares cercanos a su rango de medición deseado. Revuelva para los mejores resultados. Luego de la calibración va a ser visible en la visualización inferior la eficiencia del electrodo que corresponde a la medición activa. “- - -“ va a aparecer si no se lleva a cabo ninguna calibración. Presione MEAS para regresar al modo de medición. Rango # Rango de Conductividad r1 r2 r3 r4 r5 0.00 - 19.99 µS 20.0 - 199.9 µS 200.0 - 1999 µS 2.00 - 19.99 mS 20.0 - 200.0 mS Valores de Calibración Automáticos Temperatura de Normalización 25 ºC 20 ºC Ninguno Ninguno 84 µS 76 µS 1413 µS 1278 µS 12.88 mS 11.67 mS 111.8 mS 102.1 mS Utilizando Reconocimiento Automático de Conductividad 1. Desde el modo de medición de conductividad sumerja el sensor en su estándar – 84 μS, 1413 μS, 12,88 mS, o 111,8 mS, luego presione CAL. La visualización primaria va a buscar al estándar más cercano. La visualización secundaria va a mostrar el valor no ajustado y la visualización inferior va a mostrar la temperatura. Se permite un valor de calibración en cada rango. 2. Cuando aparece el indicador LISTO presione ENTER para aceptar. En la medición primaria va a brillar “LISTO”. 3. Para calibrar otro estándar, lave sus electrodos y luego sumérjalos en su próximo estándar o presione MEAS para regresar al modo de medición. La pantalla primaria va a buscar al valor estándar más cercano, que no ha sido calibrado aún, mientras que la pantalla secundaria va a mostrar el valor no ajustado. Cuando aparece el indicador LISTO presione ENTER para aceptar. 4. Luego de que se haya calibrado la cantidad deseada de estándares, la visualización inferior va a indicar la eficiencia en el modo de medición. Utilizando Ajuste Manual 1. Desde el modo de medición Conductividad/TDS/SALINIDAD sumerja el sensor en su estándar y luego presione CAL. 2. Cuando aparezca el indicador LISTO utilize las flechas arriba/abajo para ajustar la medición primaria para que sea igual que el valor estándar a la temperatura medida, luego presione ENTER. 3. Lave sus electrodos y sumerja el sensor en su próximo estándar o presione MEAS para regresar al modo de medición, Repita si es necesario. RLuego de que se haya calibrado la cantidad deseada de estándares, la visualización inferior va a indicar la eficiencia en el modo de medición. Calibración de temperatura/ATC Manual La calibración de la temperatura se recomiendo antes del primer uso, luego del reemplazo del sensor ATC y periódicamente según sea necesario. 1. Oprima CAL de cualquier medición, luego oprima MODE (Modo). 2. Siga con el paso 3 por una ATC manual, de otra manera sumerja el sensor de temperatura en una solución con una temperatura determinada conocida. La pantalla superior muestra la temperatura activa mientras que la pantalla inferior muestra la temperatura estándar de fábrica sin ajuste. 3. Use las flechas arriba/abajo para ajustar el monitor superior. Oprima ENTER para aceptar la temperatura de calibración. El valor ajustable máximo es ±10 ºC (o ±18 ºF) desde el valor de fábrica. Mensajes de error “ERR” va a aparecer si existe una condición de error o si es presionada la tecla equivocada. Ejemplos comunes son entre otros: • Apretar ENTER durante la calibración antes de que aparezca la indicación “READY” (Listo). Espere a que aparezca la indicación “READY” (Listo) antes de apretar ENTER. • RI (Rango inferior) • RS (Rango superior) Uso intencionado, mantenimiento y precauciones Estos medidores portables utilizan sensores para detectar varios parámetros para mediciones basadas en agua. Para el mantenimiento de rutina desconecte el cable de alimentación o la batería y luego pase un trapo húmedo por la pantalla. Si es necesario puede utilizarse agua tibia o un detergente suave basado en agua. Remueve inmediatamente cualquier substancia derramada sobre el medidor con el procedimiento de limpieza correcto para el tipo de substancia derramada. • No utilice este equipo en atmósferas potencialmente peligrosas. • Vea las instrucciones del electrodo para su uso, almacenamiento y limpieza. • Asegúrese de que ningún liquido ingrese al instrumento. • No use químicos de limpieza agresivos (solventes o substancias similares). • No hay partes en el interior que pueden ser mantenidos por el usuario. El intento de mantener partes internas puede anular la garantía. • ADVERTENCIA: No está permitido ninguna modificación de este equipo. Condiciones de funcionamiento del instrumento Temperatura de ambiente de 5 a 45 ºC funcionamiento Humedad relativa de 5 a 85 %, no condensada funcionamiento Temperatura de almacenamiento -20 a +60 ºC Humedad relativa de 5 a 85 %, no condensada almacenamiento Contaminación Grado de 2 Sobretensión Categoría de II Peso 500 g Tamaño (L x A x A) 21,15 x 9,87 x 5,85 cm Ordenanzas y Seguridad CE, TUV 3-1, FCC Clase A Potencia Nominal Entrada de CC: 9 VDC 1 A 2 x AA (LR6) 1,5 V baterías Requerimientos de batería (reemplazar baterías cuando parpadea el cartel de la batería) Vibración Envío/Manejo conforme a ISTA #1A Test de caída en envase conforme Shock a ISTA #1A Carcasa (diseñado para cumplir) IP67 (utilizando cubiertas de goma) Transformador universal Condiciones de funcionamiento Temperatura de Ambiente de 5 a 50 ºC funcionamiento Humedad relativa de 0 a 90 %, no condensada funcionamiento Temperatura de almacenamiento -20 a +75 ºC Humedad relativa de 0 a 90 %, no condensada almacenamiento Contaminación Grado de 2 Sobretensión Categoría de II I/P: 100 - 240 V, 50/60 Hz, 0.3A Potencia Nominal O/P: 9 VDC 1 A Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24, Singapore 139949 Tel: (65) 6778-6876 Fax: (65) 6778-0086 [email protected] www.eutechinst.com Oakton Instruments 625 East Bunker Court, Vernon Hills, IL, 60061, USA Tel: 1-888-462-5866 Fax: 1-847-247-2984 [email protected] www.4oakton.com 68X651902 08/14 Rev0 Part of Thermo Fisher Scientific 150 & 450 Series Waterproof Handheld Meter 150 & 450 Series Medidor de mano resistente al agua Conductivity/TDS/Salinity Operation Instructions Models: CON 150 Con/TDS CON 450 Con/TDS/Salinity PC 450 pH/mV/Con/TDS Getting Started/Connections After installing (2) AA batteries and/or connecting the optional 110/220 VAC power supply, connect the desired sensors to the corresponding ports. BNC: pH, mV/ORP or Ion Selective (PC 450 only) ATC/Con: 6-Pin ATC / Conductivity or ATC A/C Power Adapter Output (USB or RS232 – PC or Printer Communication) 12 mm and 16 mm probes can utilize the Grip-Clip™ to attach one or more sensors to a beaker and to the instrument as needed. The stand can be extended as shown above or used for wall-mounting. Keypad Functions Press once to power ON in the mode that was previously used. Press again to turn backlight on for one minute or off (450 series only). Hold for 3 seconds to power OFF. Toggle between measurement and calibration modes. In SETUP mode, BACK serves to return to the previous menu option or setting. Confirm calibration values in CAL mode. Confirm selections in SETUP mode. Freeze or release the measured reading. Customize instrument settings and preferences. (See also Setup Programs) Toggle between available measurement types. Save measurement into memory. Increase value or scroll up in SETUP or manual calibration. Recall saved values from memory. Decrease value or scroll down in SETUP or manual calibration. Send output data to printer or computer. (450 series only). Setup Programs To access the settings below, press SETUP. Up/down arrows will display the available options. Press ENTER to accept the desired setting, or BACK to return to the previous option and/or exit. Configuration Options • Ready indicator ON / OFF / or Automatic HOLD when stable • Choose ºCelsius or ºFahrenheit Conductivity Cell Options • Cell Constant. Choose (0.10 / 1.00 / 10.0) • Temperature Coefficient. Adjust from 0.00 - 10.0 %/ºC • Total Dissolved Solids Factor. Adjust from (0.00 - 1.00) • Normalization Temperature. Choose (15.0 / 20.0 / 25.0 / 30.0 °C) • Choose Single-Point Calibration (SPC) to apply a single calibration value across all ranges or Multi-Point Calibration (MPC) to calibrate individual range(s). • Choose Automatic Calibration Standards or Manual Adjustment Select Calibration Due Reminder • Set number of days from 0-60 for desired parameter View Calibration Data (for the parameter being measured) • Press ENTER to view each point that is calibrated with the associated calibration range. View Electrode Data (for the parameter being measured) • Press ENTER to view the electrode efficiency of each point that is calibrated with the associated calibration range. System Settings • Data Logging: MANUAL upon key press only TIMED interval. Choose (SEC / MIN / HOUR) interval. • Automatic shut off after 10 minutes. Choose ON or OFF. • Clock Settings: Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY). Time: Choose (24HR or 12HR). If 12HR, choose AM or PM. • Set Printer Type: CSV (Comma Separated Values) – best format for computer Printer (Text) – best format for printer. Choose Manual (MAN) upon key press or TIMED interval. If timed, choose (SEC / MIN / HOUR). Reset • NO. Exits from reset menu options without action. • FACTORY RESET. Returns all settings except date/time and ATC calibration to factory default values after ENTER is pressed then restarts meter. • DATA RESET. Erases data stored in memory while retaining other settings after ENTER is pressed. • CALIBRATION RESET. Erases non-ATC calibration data while retaining other settings after ENTER is pressed. Conductivity/TDS/Salinity Calibration For best results, periodic calibration with known accurate standards is recommended. Calibrate with standard(s) near your intended measuring range. Provide stirring for best results. After calibration, the electrode efficiency that corresponds to the active measurement will be visible on the bottom display. “- - -“ will be shown if no calibration is performed. Press MEAS to return to measurement mode at any time. Range # Conductivity Range r1 r2 r3 r4 r5 0.00 - 19.99 µS 20.0 - 199.9 µS 200.0 - 1999 µS 2.00 - 19.99 mS 20.0 - 200.0 mS Automatic Calibration Values Normalization Temperature 25 ºC 20 ºC None None 84 µS 76 µS 1413 µS 1278 µS 12.88 mS 11.67 mS 111.8 mS 102.1 mS Using Automatic Conductivity Recognition 1. From Conductivity measurement mode, dip the sensor into your standard – either 84 μS, 1413 μS, 12.88 mS, or 111.8 mS, then press CAL. The primary display will search for the nearest standard. The secondary display will show the un-adjusted value and the bottom display will show the temperature. One calibration value in each range is allowed. 2. When the READY indicator appears, press ENTER to accept. The primary reading will flash “DONE”. 3. To calibrate another standard, rinse your electrode(s) then dip into your the next standard or press MEAS to return to measurement mode at any time. The primary display will search for the nearest standard value that has not yet been calibrated, while the secondary display will show the unadjusted value. When the READY indicator appears, press ENTER to accept. 4. After the desired number of standards has been calibrated, the bottom display will show the efficiency in measurement mode. Using Manual Adjustment 1. From Conductivity/TDS/SALINITY measurement mode, dip the sensor into your standard then press CAL. Parte de Thermo Fisher Scientific Conductividad/TDS/Salinidad Instrucciones de funcionamiento 2. When the READY indicator appears use up/down arrows to adjust the primary reading to match the standard value at the measured temperature, then press ENTER. 3. Rinse your electrode(s) then dip the sensor into your next standard, or press MEAS to return to measurement mode. Repeat as necessary. After the desired number of standards have been calibrated, the bottom display will show the efficiency in measurement mode. Modelos: Temperature Calibration/Manual ATC Temperature calibration is recommended prior to first use, after ATC sensor replacement, and periodically as needed. 1. Press CAL from any measurement, then press MODE. 2. Skip to step 3 for manual ATC, otherwise, dip the temperature sensor into a solution with a known accurate temperature. The upper display shows the active temperature while the lower display shows the factory default temperature without adjustment. 3. Use up/down arrows to adjust the upper display. Press ENTER to accept the calibration temperature. The maximum adjustable value is ±10 ºC (or ±18 ºF) from factory default. Error Messages CON 150 Con/TDS CON 450 Con/TDS/Salinidad Primeros pasos/Conexiones Luego de insertar (2) baterías AA y/o conectar el suministro de energía opcional de 110/220 VAC conecte los sensores deseados a los puertos correspondientes. BNC: pH, mV/ORP o selectivo de iones (sólo PC 450) “ERR” will appear when an error condition exists or the incorrect key is pressed. Common examples include: • Pressing ENTER during calibration before the “READY” indicator appears. Wait for the “READY” indicator before pressing ENTER. • UR (Under Range) • OR (Over Range) Intended Use, Maintenance & Precautions These handheld meters use sensors to detect various parameters for water-based measurements. For routine maintenance disconnect the power cord or battery, then dust or wipe the display using a damp cloth. If necessary, warm water or a mild water based detergent can be used. Immediately remove any spilled substance from contact with the meter using the proper cleaning procedure for the type of spill. • Do not use this equipment in potentially explosive atmospheres. • Refer to the electrode instructions for use, storage and cleaning. • Ensure that no liquid enters the instrument. • Do not use any aggressive cleaning chemicals (solvents or similar agents). • There are no user serviceable parts inside. Attempts to service internal parts may void the warranty. • WARNING: No modification of this equipment is allowed. Instrument Operating Conditions Operating Ambient Temp. 5 to 45 ºC Operating Relative Humidity 5 to 85 %, non-condensing Storage Temp. -20 to +60 ºC Storage Relative Humidity 5 to 85 %, non-condensing Pollution Degree 2 Overvoltage Category II Weight 500 g Size (L x W x H) 21.15 x 9.87 x 5.85 cm Regulatory & Safety CE, TUV 3-1, FCC Class A Power Rating DC Input: 9 VDC 1 A 2 x AA (LR6) 1.5 V batteries Battery Requirement (replace batteries when battery sign blinks) Vibration Shipping/handling per ISTA #1A Drop test in packaging per Shock ISTA #1A Enclosure (Designed To Meet) IP67 (using rubber covers) Universal Power Adapter Operating Conditions Operating Ambient Temp. 0 to 50 ºC Operating Relative Humidity 0 to 90 %, non-condensing Storage Temp. -20 to +75 ºC Storage Relative Humidity 0 to 90 %, non-condensing Pollution Degree 2 Overvoltage Category II I/P: 100 - 240 V, 50/60 Hz, 0.3A Power Rating O/P: 9 VDC 1 A Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24, Singapore 139949 Tel: (65) 6778-6876 Fax: (65) 6778-0086 [email protected] www.eutechinst.com Oakton Instruments 625 East Bunker Court, Vernon Hills, IL, 60061, USA Tel: 1-888-462-5866 Fax: 1-847-247-2984 [email protected] www.4oakton.com 68X651902 08/14 Rev0 PC 450 pH/mV/Con/TDS ATC/Con: 6-Pin ATC / Conductividad o ATC Transformador A/C Salida (USB o RS232 – Comunicación de PC o Impresora) Sondas de 12 mm y 16 mm pueden utilizar Grip-Clip™ para sujetar uno o más sensores a un vaso y si es necesario al instrumento. El pie puede ser extendido como se muestra arriba o puede ser montado a la pared. Funciones del teclado Presione una vez, para encender en el modo usado previamente. Presione otra vez, para encender la retroiluminación por un minuto o apagar (sólo serie 450). Mantenga apretado por 3 segundos. Alterne ente los modos medición y calibración. En el modo CONFIGURACIÓN la tecla VOLVER sirve para volver a la opción previa del menú o ajuste. Confirmar los valores de calibración en el modo CAL. Confirmar selecciones en el modo CONFIGURACIÓN. Congelar o liberar el valor medido. Personalizar los ajustes del instrumento y las preferencias. (Vea también los programas de configuración). Alterne entre los tipos de medición disponibles. Guardar la medición en la memoria. Aumente el valor o desplace hacia arriba en CONFIGURACIÓN o calibración manual. Volver a mostrar valores guardados en la memoria. Reduzca el valor o desplace hacia abajo en CONFIGURACIÓN o calibración manual. Envíe los datos de salida a una impresora o computadora. (Sólo serie 450). Programas de configuración Para acceder a los ajustes abajo presione CONFIGURACIÓN, Las flechas arriba/abajo van a hacer aparecer las opciones disponibles. Presione ENTER para aceptar el ajuste deseado o BACK (Volver) para volver a la opción anterior y/o salir. Opciones de Configuración • Indicador listo ON / OFF / o HOLD (Mantenimiento) automático si estable • Elija ºCelsius o ºFahrenheit Opciones de Célula de Conductividad • Constante de Célula. Elija (0.10 / 1.00 / 10.0) • Coeficiente de Temperatura. Ajuste entre 0.00 - 10.0 %/ºC • Factor total de sólidos disueltos. Ajuste entre (0.00 - 1.00) • Temperatura de normalización. Seleccione (15.0 / 20.0 / 25.0 / 30.0 °C) • Seleccione Calibración de un punto (SPC) para aplicar un sólo valor de calibración a todos los rangos o Calibración de varios puntos (MPC) para calibrar rangos individuales. • Seleccione Estándares de Calibración Automáticos o Ajuste Manual Seleccione Notificación de Calibración Pendiente • Fije la cantidad de días entre 0 y 60 para el parámetro deseado Ver Datos de Calibración (para el parámetro que se está midiendo) • Presione ENTER para ver cada punto que va a ser calibrado con el rango de calibración asociado. Ver Datos de Electrodo (para el parámetro que se está midiendo) • Presione ENTER para ver la eficiencia del electrodo de cada punto que se está calibrando con su rango de calibración asociado. Ajustes del Sistema • Registro de datos: MANUAL al apretar un botón sólo intervalo TIMED (Cronometrado). Elija (SEG / MIN / HORA) intervalo. • Apagado automático después de 10 minutos. Elija ON oder OFF. • Ajustes del reloj: Fecha: Elija USA (MM/DD/AAAA) o Euro (DD/MM/AAAA). Tiempo: Elija (24HR o 12HR). Si 12HR elija AM o PM. • Ajustar tipo de impresora: CSV (Valores Separados por Coma) – mejor formato para la computadora Impresora (texto) – mejor formato para impresora. Elija Manual (MAN) al apretar un botón o intervalo TIMED (Cronometrado). Si cronometrado, elija (SEG / MIN / HORA). Resetear • NO (NRO). de salidas de las opciones del menú de reseteo sin acción. • FACTORY RESET (RESETEO DE FÁBRICA). Retorna todos los ajustes von excepción de fecha/tiempo y calibración ATC a los valores estándares de fábrica luego de oprimir ENTER el medidor es reseteado. • DATA RESET (RESETEO DE DATOS). Elimina datos guardados en la memoria mientras otros ajustes se conservan luego de oprimir ENTER. • CALIBRATION RESET (RESETEO DE CALIBRACIÓN). Elimina datos no pertenecientes a la calibración ATC mientras otros ajustes se conservan luego de oprimir ENTER. Conductividad/TDS/Salinidad Calibración Para mejores resultados se recomienda una calibración periódica con estándares conocidos y precisos. Calibre con estándares cercanos a su rango de medición deseado. Revuelva para los mejores resultados. Luego de la calibración va a ser visible en la visualización inferior la eficiencia del electrodo que corresponde a la medición activa. “- - -“ va a aparecer si no se lleva a cabo ninguna calibración. Presione MEAS para regresar al modo de medición. Rango # Rango de Conductividad r1 r2 r3 r4 r5 0.00 - 19.99 µS 20.0 - 199.9 µS 200.0 - 1999 µS 2.00 - 19.99 mS 20.0 - 200.0 mS Valores de Calibración Automáticos Temperatura de Normalización 25 ºC 20 ºC Ninguno Ninguno 84 µS 76 µS 1413 µS 1278 µS 12.88 mS 11.67 mS 111.8 mS 102.1 mS Utilizando Reconocimiento Automático de Conductividad 1. Desde el modo de medición de conductividad sumerja el sensor en su estándar – 84 μS, 1413 μS, 12,88 mS, o 111,8 mS, luego presione CAL. La visualización primaria va a buscar al estándar más cercano. La visualización secundaria va a mostrar el valor no ajustado y la visualización inferior va a mostrar la temperatura. Se permite un valor de calibración en cada rango. 2. Cuando aparece el indicador LISTO presione ENTER para aceptar. En la medición primaria va a brillar “LISTO”. 3. Para calibrar otro estándar, lave sus electrodos y luego sumérjalos en su próximo estándar o presione MEAS para regresar al modo de medición. La pantalla primaria va a buscar al valor estándar más cercano, que no ha sido calibrado aún, mientras que la pantalla secundaria va a mostrar el valor no ajustado. Cuando aparece el indicador LISTO presione ENTER para aceptar. 4. Luego de que se haya calibrado la cantidad deseada de estándares, la visualización inferior va a indicar la eficiencia en el modo de medición. Utilizando Ajuste Manual 1. Desde el modo de medición Conductividad/TDS/SALINIDAD sumerja el sensor en su estándar y luego presione CAL. 2. Cuando aparezca el indicador LISTO utilize las flechas arriba/abajo para ajustar la medición primaria para que sea igual que el valor estándar a la temperatura medida, luego presione ENTER. 3. Lave sus electrodos y sumerja el sensor en su próximo estándar o presione MEAS para regresar al modo de medición, Repita si es necesario. RLuego de que se haya calibrado la cantidad deseada de estándares, la visualización inferior va a indicar la eficiencia en el modo de medición. Calibración de temperatura/ATC Manual La calibración de la temperatura se recomiendo antes del primer uso, luego del reemplazo del sensor ATC y periódicamente según sea necesario. 1. Oprima CAL de cualquier medición, luego oprima MODE (Modo). 2. Siga con el paso 3 por una ATC manual, de otra manera sumerja el sensor de temperatura en una solución con una temperatura determinada conocida. La pantalla superior muestra la temperatura activa mientras que la pantalla inferior muestra la temperatura estándar de fábrica sin ajuste. 3. Use las flechas arriba/abajo para ajustar el monitor superior. Oprima ENTER para aceptar la temperatura de calibración. El valor ajustable máximo es ±10 ºC (o ±18 ºF) desde el valor de fábrica. Mensajes de error “ERR” va a aparecer si existe una condición de error o si es presionada la tecla equivocada. Ejemplos comunes son entre otros: • Apretar ENTER durante la calibración antes de que aparezca la indicación “READY” (Listo). Espere a que aparezca la indicación “READY” (Listo) antes de apretar ENTER. • RI (Rango inferior) • RS (Rango superior) Uso intencionado, mantenimiento y precauciones Estos medidores portables utilizan sensores para detectar varios parámetros para mediciones basadas en agua. Para el mantenimiento de rutina desconecte el cable de alimentación o la batería y luego pase un trapo húmedo por la pantalla. Si es necesario puede utilizarse agua tibia o un detergente suave basado en agua. Remueve inmediatamente cualquier substancia derramada sobre el medidor con el procedimiento de limpieza correcto para el tipo de substancia derramada. • No utilice este equipo en atmósferas potencialmente peligrosas. • Vea las instrucciones del electrodo para su uso, almacenamiento y limpieza. • Asegúrese de que ningún liquido ingrese al instrumento. • No use químicos de limpieza agresivos (solventes o substancias similares). • No hay partes en el interior que pueden ser mantenidos por el usuario. El intento de mantener partes internas puede anular la garantía. • ADVERTENCIA: No está permitido ninguna modificación de este equipo. Condiciones de funcionamiento del instrumento Temperatura de ambiente de 5 a 45 ºC funcionamiento Humedad relativa de 5 a 85 %, no condensada funcionamiento Temperatura de almacenamiento -20 a +60 ºC Humedad relativa de 5 a 85 %, no condensada almacenamiento Contaminación Grado de 2 Sobretensión Categoría de II Peso 500 g Tamaño (L x A x A) 21,15 x 9,87 x 5,85 cm Ordenanzas y Seguridad CE, TUV 3-1, FCC Clase A Potencia Nominal Entrada de CC: 9 VDC 1 A 2 x AA (LR6) 1,5 V baterías Requerimientos de batería (reemplazar baterías cuando parpadea el cartel de la batería) Vibración Envío/Manejo conforme a ISTA #1A Test de caída en envase conforme Shock a ISTA #1A Carcasa (diseñado para cumplir) IP67 (utilizando cubiertas de goma) Transformador universal Condiciones de funcionamiento Temperatura de Ambiente de 5 a 50 ºC funcionamiento Humedad relativa de 0 a 90 %, no condensada funcionamiento Temperatura de almacenamiento -20 a +75 ºC Humedad relativa de 0 a 90 %, no condensada almacenamiento Contaminación Grado de 2 Sobretensión Categoría de II I/P: 100 - 240 V, 50/60 Hz, 0.3A Potencia Nominal O/P: 9 VDC 1 A Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24, Singapore 139949 Tel: (65) 6778-6876 Fax: (65) 6778-0086 [email protected] www.eutechinst.com Oakton Instruments 625 East Bunker Court, Vernon Hills, IL, 60061, USA Tel: 1-888-462-5866 Fax: 1-847-247-2984 [email protected] www.4oakton.com 68X651902 08/14 Rev0
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

Eutech CON El manual del propietario

Categoría
Medición
Tipo
El manual del propietario