Ivoclar Vivadent Bluephase N Instructions For Use Manual

Tipo
Instructions For Use Manual
Bluephase
®
N
EN Instructions for Use
Page 2
FR Mode d’emploi
Page 16
ES Instrucciones de uso
Pagina 30
PT Instruções de Uso
Página 44
TR Kullanma Talimatı
Sayfa 58
RU àÌÒÚÛ͈Ëfl
ëÚ. 72
Appendix
Page 86
Manufacturer: Ivoclar Vivadent AG
FL-9494 Schaan/Liechtenstein
In USA distributed by
Ivoclar Vivadent Inc.,
175 Pineview Drive,
Amherst, NY 14228, USA
Rx ONLY
For dental use only!
Made in Austria
2
Dear Customer,
Optimum polymerization is an important require-
ment for all light cured materials in order to con-
sistently produce high quality restorations.
The polymerization light selected also plays a deci-
sive role in this respect. Therefore, we would like
to thank you for having purchased Bluephase N.
Bluephase N is a high-quality medical device which
has been designed according to the latest standard
of science and technology in compliance with the
relevant industry standards.
These Instructions for Use will help you safely
start-up the device, make full use of its capabili-
ties, and ensure a long service life.
Should you have any further questions, please do
not hesitate to contact us (see addresses on the
reverse page).
Your Ivoclar Vivadent Team
3
Product overview 4
List of parts
Indicators on the charging base
Indicators on the handpiece
Safety 6
Intended use
Indication
Signs and symbols
Safety notes
Start-up 8
Charging base
– Handpiece
Battery
Click & Cure corded operation
Operation 11
Selecting the curing program and the curing time
Cure Memory function
– Start
Acoustic signals
Light intensity
Measuring the light intensity
Maintenance and cleaning 13
– Housing
Light probes
Battery contacts
– Disposal
What if …? 14
Warranty / Procedure in case of repair 15
Product specifications 15
Table of Contents
English
4
List of parts
1 Handpiece
2 Start button
3 Display
4 Program selection button
5 Time selection button
6 Volume button
7 Power indicator
8 Charging base
9 Power cord
10 Power pack
11 Battery
12 Light probe 10 mm
13 Anti-glare cone
10
11
311213
2
9
7 8
5 6 4
5
Indicators on the charging base
Indicators on the handpiece
Indicator lights up in blue - the charging base is
connected to the power supply
Curing program and operating state
Selected indications and curing time
Charging status of the battery
6
Safety
Intended use
Bluephase N is an LED polymerization light that produces
energy-rich blue light. It is used for the polymerization of
light-curing dental materials immediately at the dental
unit. The intended use also includes the observation of
the notes and regulations in these Instructions for Use.
Indication
With its “Polywave
®
” broadband spectrum, Bluephase N
is suitable for the polymerization of all light-curing dental
materials curing in the wavelength range of 385–515 nm.
These materials include restoratives, bonding agents/
adhesives, bases, liners, fissure sealants, temporaries, as
well as luting materials for brackets and dental-lab
materials such as ceramic inlays.
Signs and symbols
Non permissible use
Symbols on the curing light
Double insulation
(device complies with safety class II)
Protection against electrical shock
(BF type apparatus)
Observe Instructions for Use
Observe Instructions for Use
Caution
The curing light must not be disposed of in
the normal domestic waste. Information
regarding disposal of the light can be found
on the respective national Ivoclar Vivadent
homepage.
Recyclable
Safety notes
The Bluephase N is is an electronic device and a medical
product which is subject to IEC 60601-1 (EN 60601-1)
and EMC directives, as well as the 93/42/EEC Medical
Device Directive. The appliance complies with the
relevant EU regulations and is classified as an LED
Class 2 product.
The apparatus has been shipped from the manufacturer
in a safe and technically sound condition. In order to
maintain this condition and to ensure risk-free operation,
the notes and regulations in these Instructions for Use
have to be observed. To prevent damage to equipment
and risks for patients, users, and third parties, the
following safety instructions have to be observed.
Materials, the polymerization of which is
activated outside the wavelength range of
385–515 nm (no materials known to date.)
If you are not sure about certain products,
please ask the manufacturer of the corres-
ponding material.
Do not charge or use the appliance near
flammable or combustible substances.
Portable and mobile high-frequency communi-
cation devices may interfere with medical
equipment. The use of mobile phones during
operation is not allowed.
7
Usage and liability
Bluephase N must only be employed for the intended
use. Any other uses are contraindicated. Liability can-
not be accepted for damage resulting from misuse or
failure to observe the Operating Instructions.
The user is responsible for testing Bluephase N for its
use and suitability for the intended purposes. This is
particularly important if other equipment is used in the
immediate vicinity of Bluephase N at the same time.
Use only original spare parts and accessories from
Ivoclar Vivadent (see Accessories). The manufacturer
does not accept any liability for damage resulting from
the use of other spare parts or accessories.
The light probe is an applied part and may warm up to
a maximum of 45 °C (113 °F) at the interface to the
handpiece during operation.
Operating voltage
Before switching on, make sure that
a) the voltage indicated on the rating plate complies with
the local power supply.
b) the unit has acquired the ambient temperature.
If the battery or power pack are used separately, e.g.
during start-up or Click & Cure corded operation, contact
with patients or third parties must be prevented. Do not
touch the exposed contacts of the battery or connection
plug (power pack).
Assumption of impaired safety
If it has to be assumed that safe operation is no longer
possible, the power must be disconnected and the bat-
tery removed to avoid accidental operation. This may be
the case, for example, if the apparatus is visibly dam-
aged or no longer works correctly. A complete disconnec-
tion from the power supply is only ensured when the
power cord is disconnected from the power source.
Eye protection
Direct or indirect exposure of the eyes must be prevent-
ed. Prolonged exposure to the light is unpleasant for the
eyes and may result in injury. Therefore, using the sup-
plied anti-glare cones is recommended. Individuals who
are generally sensitive to light, who take photosensitiz-
ing drugs, have undergone eye surgery, or people who
work with the apparatus or in its vicinity for long periods
of time should not be exposed to the light of this device
and wear protective goggles (orange) that absorb light
below 515 nm.
Battery
Caution: Use only original spare parts, particularly Ivoclar
Vivadent batteries and charging bases.
Do not short circuit battery. Do not store at temperatures
above 40 °C / 104 °F (or 60 °C / 140 °F for a short
period). Always store batteries charged. The storage
period must not exceed 6 months. May explode if
disposed of in fire.
Please note that the lithium polymer battery may react
with explosion, fire, and smoke development in case of
inappropriate handling and mechanical damage.
Damaged lithium polymer batteries must no longer be
used.
The electrolytes and electrolyte fumes released during
exlosion, fire, and smoke development are toxic and
corrosive. In case of contact with the eyes and skin,
immediately rinse with copious amounts of water. Avoid
the inhalation of the fumes. See a physician immediately
in case of indisposition.
Heat development
As it is the case with all high-performance lights, the
high light intensity results in a certain heat development.
Prolonged exposure of areas near the pulp and soft
tissues may result in irreversible or reversible damage.
Therefore, the recommended curing times have to be
observed. Uninterrupted curing times of more than
20 seconds on the same tooth surface, as well as direct
contact with the gingiva, oral mucous membrane, or
skin, have to be prevented. If required, polymerize
indirect restorations at intermitting intervals of
20 seconds each or use external cooling with an air
stream.
Check the delivery for completeness and any possible
transportation damage (see delivery form). If components
are missing or damaged, immediately contact your dealer
or your Service Center.
Charging base
Before you switch on the device, make sure that the volt-
age mentioned on the rating plate complies with your
local power supply. The rating plate is attached to the
underside of the charging base.
Place the charging base on a suitable, flat table top.
Remove the protective foil from the connection plug of
the power pack. Slide the connection plug of the power
pack into the socket on the underside of the charging
base. Tilt it slightly and apply slight pressure until you
hear and feel it snap into place.
Connect the power cord with the power supply and the
power pack. The Power ON indicator on the left side of
the housing lights up in blue (see Indicators on the
charging base).
Handpiece
Before you insert the light probe, remove the protective
foil on the corresponding opening of the handpiece.
Slightly turn the light probe while attaching it to the
handpiece.
After that, mount anti-glare protection on the light
probe.
8
Start-up
Battery
We recommend fully charging the battery before the first
use.
If the battery is fully charged, it features a curing
capacity of approximately 60 minutes.
Slide the battery straight into the handpiece until you
hear and feel it click into place.
Gently place the handpiece in the corresponding rest in
the charging base. If a sleeve is used, please remove it
before you charge the battery.
If possible, use the light always with a fully charged
battery. This will prolong the service life. It is therefore
recommended to place the handpiece into the charging
base after each patient. The charging time for empty
batteries is 2 hours.
Charging status of the battery
With the handpiece switched on, the current charging
status is shown on the display as follows:
Battery fully charged (curing capacity of
approximately 60 minutes)
Battery half full
Reserve (The last bar in the battery indica-
tor is red in the reserve mode. The battery
has to be charged as soon as possible.)
When the battery is being charged, „Char-
ging Battery“ briefly appears in the display
before the display returns to the stand-by
mode.
If the battery is completely empty, the
handpiece automatically switches to the
stand-by mode. The light can no longer
be called up and the curing program and
curing time can no longer be set. However,
the handpiece can be used in the Click &
Cure corded operation.
Since the battery is a consumable, it has to be replaced
after its typical life cycle has expired after 2½ years.
See battery label for the age of the battery.
9
Click & Cure corded operation
Bluephase N can be used in corded operation at any
time, but particularly when the battery is completely
empty.
For this purpose, remove the battery from the handpiece.
Then remove the power pack from the underside of the
charging base. Do not pull on the power cord.
Insert the connection plug straight into the handpiece
until you hear and feel it click into place.
During corded operation, the charging base cannot
charge the battery.
The apparatus is only completely disconnected from the
power supply if the power cord has been pulled out of
the plug-in socket.
Disinfect or autoclave contaminated surfaces of the
curing light, as well as light probes, and anti-glare cones
before each use. Furthermore, make sure that the stipu-
lated light intensity permits adequate polymerization. For
that purpose, check the light probe for contamination
and damage, as well as the light intensity at regular
intervals (e.g. with the Ivoclar Vivadent radiometer
“Bluephase Meter”).
Selecting the curing program and time
The curing program and the curing time can be
individually set. Bluephase N is equipped with the
following 3 curing programs for the different indications.
The desired curing program is selected with the program
selection buttons. The display changes accordingly (see
Indicators on the handpiece).
Upon the first start-up, the following settings have been
preset:
HIGH (High Power): 10 seconds
LOW (Low Power): 10 seconds
SOFT (Soft start): 15 seconds
The intended curing time is selected using the time
selection buttons. Users may choose between 5, 10, 15,
20 and 30 seconds.
Observe the Instructions for Use of the material applied
when selecting the curing time.
The curing recommendations for composite materials
apply to all shades and, if not mentioned otherwise in
the Instructions for Use of the relevant material, to a
maximum layer thickness of 2 mm. Generally, these
recommendations apply to situations where the emission
window of the light probe is placed directly over the
material to be polymerized. Increasing the distance
between the light source and the material will require
the curing time to be extended accordingly. For instance,
if the distance to the material is 9 mm, the effective light
output is reduced by approx. 50%. In this case, the
recommended curing time has to be doubled.
10
Operation
11
Curing programs
1) Applies to a maximum layer thickness of 2 mm and provided that the Instructions for Use of the respective material do not state any other recommendation (might be the case, e.g. with dentin shades)
2) Applies to a maximum layer thickness of 4 mm and provided that the Instructions for Use of the respective material do not state any other recommendation (might be the case, e.g. with dentin shades)
3) Applies to a maximum layer thickness of 3 mm 4) Applies to light-curing (use of the base past only) 5) Applies to dual-curing
HIGH POWER Program
Consistently high light intensity for
the polymerization of restorative and
cementation materials for direct and
indirect restorations.
SOFT START Program
Step-by-step increase of the light
intensity with reduced shrinkage
stress and reduced heat development
for the polymerization of restorative
materials.
LOW POWER Program
Reduced light intensity with reduced
heat development for the polymeriza-
tion of adhesives, liners, and restora-
tive materials in areas near the pulp
when restoring Class V cavities.
650 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0 5
650 mW/cm
2
Filling materials
Composite
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
All conventional composites
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Compomer
3)
Compoglass F / Compoglass Flow
Indirect restorations / Luting materials
Variolink II Base
4)
/ Variolink Veneer / Variolink N Base
4)
/ Variolink N Clear Veneer / Variolink Ultra
5)
/
Dual Cement
5)
/ Variolink II
5)
Miscellaneous
Helioseal / Helioseal F / Helioseal Clear
Monopaque
MultiCore Flow / Multicore HB
Heliosit Orthodontic
Telio Add-On Flow
Telio Stains
IPS Empress Direct Color
IPS Empress Direct Opaque
Curing time
10 s
15 s
10 s
20 s
per mm ceramic:
10 s per segment
10 s
20 s
20 s
10 s
15 s
10 s
10 s
20 s
Adhesive
AdheSE / AdheSE One F
ExciTE / ExciTE DSC
Tetric N-Bond / Tetric N-Bond Self-Etch
Heliobond (Syntac)
Temporary materials
Telio CS Inlay/Onlay
Systemp.inlay/onlay
Fermit / Fermit N
Telio CS Link / Systemp.link
Miscellaneous
Heliosit Orthodontic
Vivaglass Liner
Curing time
10 s
10 s
10 s
10 s
20 s per segment
20 s
20 s
Filling materials
Composite
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
All conventional composites
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Compomer
3)
Compoglass F / Compoglass Flow
Curing time
15 s
20 s
15 s
20 s
Light intensity
Exposure time for composites
Exposure time for Tetric EvoCeram / IPS Empress Direct
1,200 mW/cm
2
15 s
10 s
12
Cure Memory function
The last settings used, together with the combination of
curing program and curing time, are automatically saved.
Start
The light is switched on by means of the start button.
Once the selected curing time has elapsed, the curing
program is automatically terminated. If desired, the light
can be switched off before the set curing time has
elapsed by pressing the start button again. The fan is
activated simultaneously to the light. Once the curing
time has elapsed, the fan continues to run for a certain
time to cool the apparatus. The battery must not be
removed as long as the fan is still running.
Acoustic signals
Acoustic signals can be heard for the following functions:
Start (Stop)
Every 10 seconds
Program change
Curing time change
Connecting the battery to the charging base
Inserting battery
Error message
If desired, the volume of the acoustic signals can be
adjusted. For that purpose, press the blue volume button
while the light is off and reduce the volume or turn the
acoustic signals off with the left program selection but-
ton. If you want to turn the acoustic signals back on or
increase the volume, press the blue volume button and
then the right program selection button.
Light intensity
The light intensity is maintained at a consistent level
during operation. If the supplied 10 mm light probe is
used, the light intensity has been calibrated to
1200 mW/cm
2
± 10%.
The use of a light probe other than the one provided has
a direct influence on the light intensity emitted.
In parallel-walled light probes (10 mm), the diameter is
equal at both ends. In focussing light probes (10>8 mm
light probe, Pin-Point light probe 6>2 mm), the diameter
of the rear end is larger than that of the light emission
window. The incident blue light is thus bundled to a
smaller surface, which increases the light intensity emitted.
Pin-Point light probes are suitable for the polymerization
of confined areas, such as the attachment of veneers prior
to excess removal. For thorough curing, it is necessary to
change the light probe.
Measuring the light intensity
The light intensity of the Bluephase N and the enclosed
10 mm light probe can be checked by means of the
Bluephase Meter.
If the measured value does not correspond with the
expected light intensity, proceed as follows:
Check the selected curing program.
Clean possibly dirty light sensor (radiometer).
Remove the light probe and clean the light emission
window of the handpiece with a cotton swab dipped
in alcohol.
Clean possibly dirty light probe
(see Maintenance and cleaning).
Replace a damaged light probe with a new one.
If these measures do not improve the results, please
contact your dealer or your local Service Center.
For reasons of hygiene, we recommend using a disposable
protective sleeve for each pa-tient. Make sure to fit the
protective sleeve closely to the light probe. Disinfect
contaminated surfaces of the device and anti-glare cones
(FD 366/Dürr Dental, Incidin Liquid/Ecolab) and sterilize
the light probe before each use if disposable protective
sleeves are not used. Make sure that no liquids or other
foreign substances enter the handpiece, charging base
and particularly the power pack during cleaning (risk of
electrical shock). Disconnect the charging base from the
power source when cleaning it.
Housing
Wipe the handpiece and handpiece holder with a
customary aldehyde-free disinfecting solution. Do not
clean with highly aggressive disinfecting solutions (e.g.
solutions based on orange oil or with an ethanol content
of more than 40%), solvents (e.g. acetone), or pointed
instruments, which may damage or scratch the plastic.
Clean dirty plastic parts with a soapy solution.
Light probe
Before cleaning and/or disinfecting the light probe,
pretreat it. This applies to both automated and manual
cleaning and disinfection.
Pretreatment
Remove gross contamination immediately after use or
within 2 hours. For this purpose, thoroughly rinse the
light probe under running water (for at least
10 seconds). Alternatively, use a suitable aldehyde-free
disinfecting solution to prevent the fixation of blood.
To remove contamination manually, use a soft brush
or soft cloth. Partially polymerized composite can be
removed with alcohol and a plastic spatula, if necessary.
Do not use sharp or pointed objects, as they may
scratch the surface.
Cleaning and disinfection
For cleaning, immerse the light probe in a cleaning solu-
tion and ensure that it is sufficiently covered with liquid
(ultrasound or careful brushing with a soft brush may
support the effect). A neutral-enzymatic cleaning agent is
recommended. When cleaning and disinfecting, please
make sure that the agents used are free of:
organic, mineral and oxidizing acids (the minimum
admissible pH value is 5.5)
alkaline solution (the maximum admissible pH value
is 8.5)
oxidizing agent (e.g. hydrogen peroxide)
Afterwards, remove the light probe from the solution and
thoroughly rinse it under running water (for at least
10 seconds). Cleaning in a thermal disinfector is an
effective alternative.
Sterilization
Thorough cleaning and disinfecting is imperative to
ensure that the subsequent sterilization is effective. Use
only autoclave sterilization for this purpose. The steriliza-
tion time (exposure time at sterilization temperature) is
4 minutes at 134 °C (273 °F); pressure should be 2 bar
(29 psi). Dry the sterilized light probe using either the
special drying program of your steam autoclave or hot
air. The light probe has been tested for up to 200 sterili-
zation cycles.
After that, check the light probe for damage. Hold it
against light. If individual segments appear black, glass
fibres are broken. If this is the case, replace the light
probe with a new one.
Battery contacts
To ensure reliable conductivity at all times, keep the
battery contacts free from possible contamination (e.g.
composite residue). For this purpose, clean the affected
contacts regularly in the course of the usual wipe
disinfection (after each patient).
Disposal
The curing light must not be disposed of as
urban waste. Dispose unserviceable batteries
and polymerization lights according to the
corresponding legal requirements in your
country. Batteries must not be incinerated.
13
Maintenance and cleaning
14
What if....?
Symbol Causes Error Rectification
Electronic defect in the handpiece
or battery
Apparatus is overheated or undercooled
Electronic defect in the battery
Battery empty
Charging contacts contaminated
Battery in safety mode
(integrated protective circuit)
Power pack not connected or defective
Remove and reinsert the battery. If the error remains,
replace the battery with the power pack (Click & Cure).
If the error remains, please contact your dealer or your
local Service Center.
Allow the apparatus to cool down (or assume room
temperature if it is undercooled) and try again after a
certain time. If the error remains, please contact your
dealer or your local Service Center.
Remove and reinsert the battery. If the error remains,
place the apparatus into the charging base.
If the error remains nonetheless, replace the battery
with the power pack (Click & Cure).
Please contact your dealer or your local Service Center.
Position the handpiece in the charging base. „Charg-
ing Battery“ briefly appears in the display before the
display returns to the stand-by mode. If the battery is
not charged, the contacts have to be cleaned. If the
error remains, please contact your dealer or your local
Service Center.
Check if the handpiece has been correctly placed into
the charging base. Clean battery contacts. Information:
“Charging Battery“ briefly appears in the display
before the display returns to the stand-by mode.
If cleaning the battery contacts does not improve the
results, remove the battery from the apparatus. Charge
the battery separately from the handpiece in the
charging base for approx. 10 minutes.
If the error remains, please contact your dealer or your
local Service Center.
Check if the power pack is correctly positioned in the
charging base (also check the charging contacts) or if
the power pack is connected to the power supply by
means of the power cord (display on the power pack
lights up in green if it works correctly).
If the error remains, please contact your dealer or your
local Service Center.
No display
during
charging
LED of the
charging
base is not
illuminated
15
Warranty /
Procedure in case of repair
Product specifications
The warranty period for Bluephase N is 3 years from the
date of purchase (battery 1 year).
Malfunctions resulting from faulty material or
manufacturing errors are repaired free of charge during
the warranty period. The warranty does not provide the
right to recover any material or non-material damage
other than the ones mentioned. The apparatus must only
be used for the intended purposes. Any other uses are
contraindicated. The manufacturer does not accept any
liability resulting from misuse and warranty claims
cannot be accepted in such cases. This is particularly
valid for:
damage resulting from improper handling, especially
incorrectly stored batteries (see Technical data:
Transportation and storage conditions).
damage to components resulting from wear under
standard operating conditions (e.g. battery).
damage resulting from external influences, e.g. blows,
drop to the floor.
damage resulting from incorrect set-up or installation.
damage resulting from connecting the unit to a power
supply, the voltage and frequency of which do not
comply with the ones stated on the rating plate.
damage resulting from improper repairs or modifica-
tions that have not been carried out by certified
Service Centers.
In case of a claim under warranty, the complete
apparatus (handpiece, charging base, power cord, and
power pack) must be returned, carriage paid, to the
dealer or directly to Ivoclar Vivadent, together with the
purchase document. Use the original packaging with the
corresponding cardboard inserts for transportation.
Repair work may only be carried out by a certified
Ivoclar Vivadent Service Center. In case of a defect that
cannot be rectified, please contact your dealer or your
local Service Center (see addresses on the reverse side).
A clear description of the defect or the conditions under
which the defect occurred will facilitate locating the
problem. Please enclose this description when returning
the apparatus.
Delivery form
1 Charging base, 1 Power cord, 1 Power pack, 1 Handpiece, 1 Battery
(Li-Po), 1 10 mm light probe, 3 Anti-glare cones, 1 Set of Instructions
for Use
Accessories
The following accessories are available for Bluephase N:
REF Description
656196 Light probe 10 mm, black (N)
608538 Pin-point light probe 6>2 mm, black
551756 Anti-glare cones
592496 Anti-glare shield
659956 Bluephase N battery
659681 Bluephase N handpiece
607922 Bluephase Meter
608554 Bluephase sleeves
Technical data
Operating voltage charging base 5 VDC
Operating voltage handpiece 3.7 VDC with battery
5 VDC with power pack
Power pack 100–240 VAC / 50-60 Hz /
max. 0.4 A
Manufacturer: Friwo
Output 5 VDC
TYPE 15.2630
Operating conditions
Temperature +10 °C to +35 °C (+50 to +95 °F)
Relative humidity 30% to 75%
Ambient pressure 700 hPa to 1060 hPa
Dimensions of the charging base L=205mm W=150mm H=85mm
Weight of the charging base 250 g
Charging time approx. 2 h (with the battery empty)
Power supply of the handpiece Li-Po battery
Max. battery time approx. 60 min. (with a new, fully
charged battery)
Light source Polywave
®
LED
Wavelength range 385 – 515 nm
Light intensity max. 1,200 mW/cm
2
± 10 %
Operating 5 min. on / 6 min. off
(intermittently)
Dimensions of the handpiece: L=260mm W=42mm H=120 mm
Weight of the handpiece (incl. battery) 225 g
Transportation and storage conditions:
Temperature -20 °C to +60 °C (-4 to 140 °F)
Relative humidity 10% to 75%
Ambient pressure 500 hPa to 1060 hPa
Bluephase N has to be stored in closed, roofed rooms.
Protect the device from severe jarring.
Battery
Do not store at temperatures above 40 °C / 104 °F
(or 60 °C / 140 °F for a short period).
Recommended storage temperature 15 – 30 °C (59 – 86 °F).
Always store the battery charged
and not for longer than 6 months.
16
Cher client,
Une polymérisation optimale est une condition
importante pour tous les matériaux photopoly-
mérisables de manière à produire de façon
constante et reproductible des restaurations de
haute qualité. La lampe à polymériser choisie a,
elle aussi, un rôle décisif de ce point de vue.
C‘est pour cela que nous tenons à vous remercier
d‘avoir acheté la lampe Bluephase N.
La lampe Bluephase N est un dispositif médical de
haute qualité correspondant aux dernières normes
scientifiques et technologiques ainsi qu‘aux
normes industrielles.
Ce mode d‘emploi va vous permettre de mettre en
marche votre appareil en toute sécurité, de profiter
de toutes ses possibilités, et de lui assurer une
longue durée de vie. Pour une identification facile
des composants individuels, la vue générale du
produit est présentée sous la forme d‘une carte.
Pour toute question supplémentaire,
nous vous remercions de nous contacter
(adresse sur le verso).
Votre équipe Ivoclar Vivadent
17
Vue générale 18
Listes des pièces
Indicateurs sur la base de chargement
Indicateurs sur la pièce à main
Sécurité 20
Usage habituel
Indication
Signes et symboles
Notes de sécurité
Première mise en marche 22
Base de chargement
Pièce à main
Batterie
Cordon Click & Cure
Utilisation 24
Sélection du programme et du temps de polymérisation
Fonction de mémoire de programme de polymérisation
Démarrage
Signaux acoustiques
Intensité lumineuse
Mesurer l’intensité lumineuse
Maintenance et nettoyage 27
– Logement
Conducteurs de lumiére
Contacts de la batterie
Elimination des pièces usagées
Que faire en cas de …? 28
Garantie / Procédure en cas de réparation 29
Spécifications 29
Sommaire
Francais
18
Liste des pièces détachées
1 Pièce à main
2 Bouton de mise marche
3 Affichage
4 Bouton de sélection des programmes
5 Bouton de sélection du temps
6 Bouton de réglage du volume
7 Indicateur de puissance
8 Base de chargement
9 Cordon d‘alimentation
10 Power pack
11 Batterie
12 Conducteur de lumière 10 mm
13 Embout de protection oculaire
10
11
311213
2
9
7 8
5 6 4
19
Indicateurs sur la base de chargement
Indicateurs sur la pièce à main
Les voyants s‘allument en bleu – la base de chargement
est reliée au courant électrique.
Programme de polymérisation, temps de polymérisation,
état de marche
Indications spécifiques du programme de polymérisation
sélectionné
Témoin de charge de la batterie
20
Sécurité
Usage habituel
Bluephase N est une lampe à photopolymériser LED de
haute performance qui produit une lumière bleue dense
de haute énergie. Elle est utilisée pour la polymérisation
de matériaux dentaires au fauteuil. L‘utilisation requise
inclut aussi l‘observation des recommandations et régle-
mentations du mode d‘emploi.
Indications
Grâce à son spectre à large longueur d‘ondes, Bluephase N
est particulièrement recommandée pour la polymérisa-
tion de matériaux dentaires photopolymérisables dans
une longueur d‘onde de 385–515 nm. Ces matériaux
incluent les matériaux de restauration, les bondings/
adhésifs, les bases, les liners, les scellements de sillons et
fissures, les matériaux provisoires ainsi que les colles
pour brackets orthodontiques, les matériaux de
laboratoire dentaire tels que les inlays céramiques.
Signes et symboles
Contre-indication
Symboles sur la lampe à photopolymériser
Double protection (l‘appareil répond aux
normes de sécurité de classe II)
Protection contre les décharges électriques
(type d‘appareil BF)
Respecter le mode d‘emploi
Respecter le mode d‘emploi
Mise en garde
Ne pas jeter la lampe avec les ordures
ménagères. Concernant l‘élimintation de
votre appareil, veuillez contacter votre
distributeur habituel de produits Ivoclar
Vivadent.
Recyclable
Notes de sécurité
Bluephase N est un appareil médical électronique
fabriqué conformément à la norme IEC 60601-1 (EN
60601-1 et répondant aux Directives CEM, ainsi qu‘aux
Directives des Dispositifs Médicaux 93/42/EEC. L’appareil
est conforme aux normes européennes en vigueur et il
est répertorié comme un produit LED de Classe 2.
La lampe est expédiée par le fabricant dans des
conditions de sécurité et des conditions techniques
optimales. Pour maintenir ces conditions et assurer des
opérations sans risque, il est nécessaire de respecter les
recommandations et les réglementations du mode
d‘emploi. Pour prévenir les dommages sur l‘appareil et
tout risque pour les patients, les utilisateurs et les tiers,
les instructions de sécurité suivantes doivent être
respectées.
Ne pas utiliser l‘appareil pour photopoly-
mériser les matériaux dont la polymérisation
est activée dans une longueur d‘onde en
dehors de 385–515 nm (aucun matériau
connu à ce jour). Si vous avez un doute sur
certains produits, contacter le fabricant du
matériau.
Ne pas charger ou utiliser cet appareil à
proximité de substances inflammables ou
combustibles.
Les téléphones portables et autres appareils
de communication HF (Haute Fréquence)
peuvent interférer les équipements médicaux.
L‘utilisation d‘un téléphone portable pendant
le fonctionnement de la lampe est interdite.
21
Manipulation et responsabilité
Bluephase N ne doit être utilisée que pour les usages
recommandés. Toute autre utilisation est contre-
indiquée. Nous rejetons toute responsabilité pour des
dommages liés à un usage inadapté ou une panne
pour non-respect du présent mode d‘emploi.
De plus, l‘utilisateur est responsable de l‘usage de la
lampe Bluephase N de façon appropriée en respectant
les indications recommandées, et ce particulièrement
si d‘autres équipements sont utilisés à proximité
pendant le fonctionnement de la lampe.
N‘utiliser que les pièces détachées et accessoires
fournis par Ivoclar Vivadent (voir Accessoires). Nous
rejetons toute responsabilité pour les dommages
causés par l‘utilisation d‘autres pièces détachées ou
accessoires.
L’embout lumineux est une pièce appliquée et,
pendant le fonctionnement, peut chauffer jusqu’à
45 °C maximum au niveau de l’interface avec la pièce
à main.
Voltage
Avant la mise en marche, assurez-vous que :
a) le voltage indiqué sur la base est conforme à l‘énergie
délivrée par l‘alimentation électrique.
et
b) l‘appareil se trouve à température ambiante.
Si la batterie ou le cordon d‘alimentation sont utilisés
séparément, par exemple lors de la mise en marche ou
lors de l‘opération Click & Cure raccordée au courant, les
contacts avec les patients ou les tiers doivent être évités.
Ne pas toucher les contacts accessibles de la batterie ou
les connexions électriques (power pack).
Conditions de sécurité
Si un usage en toute sécurité ne peut être garanti, la
prise de courant doit être débranchée et la batterie
retirée pour éviter tout fonctionnement accidentel. Ceci
peut être le cas, si l‘appareil est visiblement endommagé
ou ne fonctionne plus correctement. Ne pas toucher les
connexions exposées des batteries ou du cordon
d‘alimentation. La déconnexion complète n‘est assurée
que lorsque le cordon d‘alimentation est débranché.
Protection oculaire
Eviter toute exposition directe ou indirecte avec les yeux.
Une exposition prolongée à la lumière peut être inconfor-
table et provoquer des dommages. C‘est pour cela, que
l‘utilisation des embouts de protection oculaire est
recommandée. Les personnes sensibles à la lumière, qui
prennent des médicaments photosensibles ou qui ont
subi une chirurgie oculaire, les personnes qui travaillent
avec l‘appareil ou à proximité pendant de longues
périodes ne doivent pas être exposées à la lumière de
cette lampe et doivent porter des lunettes de protection
qui absorbent la lumière en dessous de 515 nm.
Batterie
Attention : N‘utiliser que les pièces d‘origine. En particulier,
n‘utiliser que les batteries et bases de chargement
d‘origine Ivoclar Vivadent. Ne pas mettre la batterie en
court-circuit. Ne pas stocker la batterie à une température
supérieure à 60°C, ni pendant plus de 6 mois sans être
rechargée. Ne pas brûler (risque d‘explosion).
Notez que toute flamme, explosion ou dégagement
de fumée dus à une mauvaise manipulation ou à un
problème mécanique peut provoquer une réaction de la
batterie Lithium Polymère. Les batteries Lithium Polymère
endommagées ne doivent plus être utilisées.
Les electrolytes et les fumées d‘électrolytes libérés par
une explosion, une flamme ou un dégagement de fumée
sont toxiques et corrosifs. En cas de contact avec les yeux
ou la peau, rincer immédiatement et abondamment à
l‘eau. Eviter d‘inhaler les fumées. Consulter rapidement
votre médecin en cas d‘indisposition.
Dégagement de chaleur
Comme cela est le cas avec toutes les lampes à haute
performance, une haute énergie lumineuse a pour
résultat un certain dégagement de chaleur. Une exposi-
tion prolongée sur la pulpe ou sur les tissus mous peut
engendrer des dommages. Il est donc conseillé de
respecter les recommandations de polymérisation liées
au programme et au temps de polymérisation. De plus,
une polymérisation continue de plus de 20 secondes sur
la même zone, ainsi qu‘un contact direct sur la gencive,
les muqueuses, ou la peau, doivent être évités. Si
nécessaire, polymériser les restaurations indirectes par
intervalles de 20 secondes ou utiliser un jet d‘air qui
permettra le refroidissement.
22
Mise en marche
Contrôler la livraison afin de vérifier qu‘il ne manque rien
et qu‘aucun dommage n‘est survenu lors du transport
(voir livraison). Si des composants manquent ou sont
endommagés, contacter immédiatement votre revendeur
ou centre de service.
Base de chargement
Avant de mettre en marche l‘appareil, assurez-vous que
le voltage inscrit sur la base de chargement est conforme
à votre alimentation électrique. Les indications sont
portées sur la plaque d‘identification qui se trouve sous
la base de chargement.
Poser la base de chargement sur une surface plate et
stable. Mettre en place la prise de connexion au niveau
du compartiment sous la base de chargement. L‘incliner
légèrement et exercer une légère pression jusqu‘à sentir
et entendre un clic.
Connecter le cordon d‘alimentation à la prise de courant.
L‘indicateur Power On sur le côté gauche du capot de la
base de chargement s‘allume en vert (voir “Indicateurs
sur la base de chargement”).
Pièce à main
Avant d‘insérer la fibre de verre conductrice de lumière,
retirer le film de protection sur l‘ouverture de la pièce à
main.
Tourner légèrement le conducteur de lumière dès que
celui-ci est fixé sur la pièce à main, puis positionner
l‘embout de protection oculaire sur le conducteur de
lumière.
23
Batterie
Il est conseillé d‘effectuer une charge complète de la
batterie avant la première utilisation.
Une batterie complètement chargée a une capacité de
polymérisation d‘environ 60 minutes.
Faire glisser la batterie dans la pièce à main jusqu‘à
l‘obtention du clic de mise en place.
Reposer délicatement la pièce à main sur la base de
chargement, à l‘emplacement prévu à cet effet. Si vous
utilisez une housse de protection, n‘oubliez pas de la
retirer avant de mettre la batterie en charge.
Si possible, toujours utiliser la lampe avec une batterie
complètement chargée, afin d‘en prolonger la durée de
vie. Il est conseillé de replacer la pièce à main sur la base
de chargement après chaque patient. Le temps de charge
d‘une batterie vide est de 2 heures.
Etat de charge de la batterie
Lorsque la pièce à main est allumée, l‘état de la charge
est affiché sur l‘écran d‘affichage comme suit
(voir “Indicateurs sur la pièce à main”) :
Batterie complètement chargée (capacité
de polymérisation d‘environ 60 minutes)
Batterie à moitié chargée
Réserve (la dernière barre de l‘indicateur de
la batterie devient rouge en mode reserve.
La batterie doit être mise en charge dès
que possible).
Lorsque la batterie est mise en charge,
„Charging Battery“ s‘affiche brièvement
avant le retour en mode stand-by de
l‘écran.
Batterie vide
Si la batterie est complètement vide, la
pièce à main se met automatiquement en
mode stand-by. La lumière ne s‘allume
plus, le programme de polymérisation et le
temps de polymérisation ne peuvent être
réglés. Cependant, la pièce à main peut
être utilisée avec le cordon Click & Cure.
La batterie est un consommable et doit, de ce fait, être
remplacée environ tous les 2½ ans, cycle de vie d‘une
batterie. Voir l‘âge de la batterie sur l‘étiquette.
24
Cordon Click & Cure
La pièce à main Bluephase N peut être branchée au
courant à n‘importe quel moment, plus particulièrement
en cas d‘urgence alors que la batterie est vide.
Pour cela, retirer la batterie de la pièce à main. Puis
retirer le cordon de branchement „power pack“ de son
logement sous le socle de la base de chargement. Ne pas
tirer sur le cordon.
Insérer le cordon de branchement “power pack” dans la
pièce à main jusqu‘à sentir et entendre un clic.
Pendant que la pièce à main est connectée au courant, la
base de chargement ne peut charger la batterie.
La déconnexion complète n‘est assurée que lorsque le
cordon d‘alimentation est débranché.
Désinfecter ou passer à l‘autoclave les différentes sur-
faces de la lampe ainsi que les conducteurs de lumière et
l‘embout de protection oculaire avant chaque utilisation.
Par ailleurs, assurez-vous que l‘intensité lumineuse
obtenue permet une polymérisation adéquate. Pour cela,
vérifier que le conducteur de lumière n‘est ni souillé ni
endommagé, puis contrôler l‘intensité régulièrement
(par ex. avec le radiomètre “Bluephase Meter”
d‘Ivoclar Vivadent).
Sélection du programme et du temps de
polymérisation
Le programme et le temps de polymérisation peuvent
être programmés individuellement. Bluephase N est
équipée des 3 programmes suivants pour les différentes
indications. Le programme de polymérisation souhaité
est sélectionné à l‘aide du bouton de sélection du
programme. L‘affichage change en conséquence (voir
“Indicateurs sur la pièce à main”).
Les programmes suivants sont pré-enregistrés :
HIGH (Haute énergie) : 10 secondes
LOW (Basse énergie) : 10 secondes
SOFT (Polymérisation progressive) : 15 secondes
Le temps de polymérisation souhaité est sélectionné à
l‘aide du bouton de sélection du temps. Les utilisateurs
ont le choix entre 5, 10, 15, 20 et 30 secondes.
Respecter le mode d‘emploi des matériaux utilisés lors de
la sélection du programme.
Les recommandations de polymérisation des matériaux
composites s‘appliquent à toutes les teintes. Sauf indica-
tion contraire dans le mode d‘emploi du matériau utilisé,
photopolymériser par couche de 2 mm maximum.
Généralement, ces recommandations s‘appliquent dans
les situations où le conducteur de lumière est orienté
directement sur le matériau à polymériser. Augmenter la
distance entre la source de lumière et le matériau
demande d‘augmenter le temps de polymérisation en
conséquence. Par exemple, si la distance avec le
matériau est de 5 mm, l‘intensité lumineuse réelle est
réduite d‘environ 50%. Dans ce cas, il convient de
doubler le temps de polymérisation.
Utilisation
25
Programmes de polymérisation
1) Valable pour une épaisseur de 2 mm maximum et dans la mesure où le mode d‘emploi du
matériau utilisé n‘indique pas d‘autre recommandation (comme cela peut être le cas par
exemple avec les teintes Dentine)
2) Valable pour une épaisseur de 4 mm maximum et dans la mesure où le mode d‘emploi du
matériau utilisé n‘indique pas d‘autre recommandation (comme cela peut être le cas par
exemple avec les teintes Dentine)
3) Valable pour une épaisseur de 3 mm maximum
4) Valable pour la photopolymérisation (utilisation de la pâte base uniquement)
5) Valable pour la polymérisation duale
HIGH POWER (Haute énergie)
Haute énergie régulière pour la
polymérisation de matériaux de
restauration et de collage dans le
cas de restaurations directes et
indirectes.
SOFT START
Augmentation progressive de
l‘intensité lumineuse avec une
réduction du stress de polymérisa-
tion et de la température pour la
polymérisation des matériaux de
restauration.
LOW POWER
Basse énergie, élévation de
température réduite pour la polyméri-
sation des adhésifs, des liners et des
matériaux de restauration pour les
cavités de classe V dans les zones
proches de la pulpe.
650 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0 5
650 mW/cm
2
Matériaux de restaurations
Composite
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
tous les composites conventionnels
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Compomère
3)
Compoglass F / Compoglass Flow
Restaurations indirectes /
Colles
Variolink II Base
4)
/ Variolink Veneer / Variolink N Base
4)
/ Variolink N Clear Veneer / Variolink Ultra
5)
/
Dual Cement
5)
/ Variolink II
5)
Divers
Helioseal / Helioseal F / Helioseal Clear
Monopaque
MultiCore Flow / Multicore HB
Heliosit Orthodontic
Telio Add-On Flow
Telio Stains
IPS Empress Direct Color
IPS Empress Direct Opaque
Temps de polymérisation
10 secondes
15 secondes
10 secondes
20 secondes
Par mm de céramique :
10 secondes par face
10 secondes
20 secondes
20 secondes
10 secondes
15 secondes
10 secondes
10 secondes
20 secondes
Adhesifs
AdheSE / AdheSE One F
ExciTE / ExciTE DSC
Tetric N-Bond / Tetric N-Bond Self-Etch
Heliobond (Syntac)
Matériaux temporaires
Telio CS Inlay/Onlay
Systemp.inlay/onlay
Fermit / Fermit N
Telio CS Link / Systemp.link
Divers
Heliosit Orthodontic
Vivaglass Liner
Temps de polymérisation
10 secondes
10 secondes
10 secondes
10 secondes
20 secondes par face
20 secondes
20 secondes
Matériaux de restaurations
Composite
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
tous les composites conventionnels
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Compomére
3)
Compoglass F / Compoglass Flow
Temps de polymérisation
15 secondes
20 secondes
15 secondes
20 secondes
Intensité
Temps d‘exposition pour les composites
Temps d‘exposition pour Tetric EvoCeram / IPS Empress Direct
1'200 mW/cm
2
15 secondes
10 secondes
26
Fonction mémoire de polymérisation
Les derniers réglages utilisés, programme de polymérisa-
tion et temps de polymérisation, sont automatiquement
mémorisés.
Démarrage
La lampe est allumée avec le bouton de démarrage. Une
fois le temps de polymérisation écoulé, le programme de
polymérisation s‘arrête automatiquement. La polymérisa-
tion peut être interrompue avant la fin du temps complet
en appuyant sur le bouton de démarrage. Le ventilateur
se déclenche en même temps que la lumière. Une fois la
polymérisation terminée, le ventilateur continue à
fonctionner un certain temps pour refroidir la lampe. La
batterie ne doit pas être retirée tant que le ventilateur
continue à fonctionner.
Signaux acoustiques
Des signaux acoustiques sont émis lors des fonctions
suivantes :
Démarrage (Arrêt)
Toutes les 10 secondes
Changement de programme
Changement du temps de polymérisation
Connexion de la batterie à la base de chargement
Insertion de la batterie
Message d‘erreur
Le volume des signaux acoustiques peut être réglé. Pour
cela, presser le bouton de volume bleu lorsque la lampe
est éteinte et réduire ou couper le volume à l‘aide du
bouton gauche de sélection du programme. Pour
réactiver les signaux acoustiques ou augmenter le
volume, presser le bouton de volume bleu, puis le bouton
droit de sélection du programme.
Intensité lumineuse
L‘intensité lumineuse est maintenue à un niveau
cons-tant pendant l‘utilisation. Si le conducteur de
lumière 10mm fourni avec la lampe est utilisé, l‘intensité
lumineuse est d‘environ 1200 mW/cm
2
± 10%.
L‘utilisation d‘embouts lumineux autres que ceux fournis
a une influence directe sur l‘intensité lumineuse émise.
Dans les embouts lumineux standards aux parois
parallèles (10 mm), le dia-mètre est égal aux deux
extrémités. Dans les embouts lumineux turbo rétrécis à
leur extrémité (10>8 mm, 6>2 mm Pin-Point), le dia-
mètre de la partie arrière est plus large que celui de la
fenêtre d‘émission de lumière. La lumière bleue incidente
est ainsi concentrée sur une plus petite surface, ce qui
augmente l‘intensité lumineuse émise.
Les embouts lumineux Pin-Point conviennent pour la
polymérisation de petites surfaces, comme par ex. la
fixation de facettes avant l‘élimination des excès. Pour
des polymérisations de zones plus larges, il est nécessaire
de changer l‘embout lumineux.
Mesurer l‘intensité lumineuse
L‘intensité lumineuse de la Bluephase N et son conduc-
teur de lumière 10 mm peuvent être contrôlés à l‘aide du
radiomètre Bluephase Meter.
Si la valeur mesurée ne correspond pas à l‘intensité
lumineuse attendue, procéder de la façon suivante :
Contrôler le programme de polymérisation sélectionné
Nettoyer le radiomètre avec un chiffon humide
Retirer le conducteur de lumière et nettoyer la fenêtre
d‘émission de lumière de la pièce à main à l‘aide d‘un
coton imbibé d‘alcool.
Nettoyer si possible les conducteurs de lumières
souillés (voir “Maintenance et nettoyage”)
Remplacer tout conducteur de lumière endommagé par
un nouveau
Si ces mesures n‘améliorent pas le résultat, veuillez
contacter votre distributeur ou le service après-vente
Ivoclar Vivadent.
27
Maintenance et nettoyage
Pour des questions d’hygiène, il est recommandé d’utiliser
une gaine de protection jetable pour chaque patient. Les
surfaces contaminées du dispositif ainsi que les embouts de
protection oculaire doivent être désinfectés (par exemple
FD333, FD366/Dürr Dental, Incidin Liquid/ Ecolab) et
l’embout lumineux stérilisé avant chaque utilisation si le
dispositif est utilisé sans une gaine de protection jetable.
Assurez-vous qu‘aucun liquide ni autre substance ne pénètre
dans la pièce à main ou la base de chargement pendant le
nettoyage (risque d‘électrocution). Déconnecter la base de
chargement du courant lors des opérations de nettoyage.
Entourage plastique
Essuyer la pièce à main avec une solution sans aldéhyde.
Ne pas utiliser de solutions corrosives (ex. solutions à
base d‘essence d‘orange ou contenant plus de 40%
d‘éthanol), de solvants (ex. l‘acétone), ou d‘instruments
pointus, qui risqueraient d‘endommager ou d‘abîmer le
plastique. Nettoyer le plastique souillé avec une solution
savonneuse.
Conducteur de lumière
Avant de nettoyer et/ou de désinfecter le conducteur de
lumière, celui-ci doit subir un traitement préalable. Ceci est
valable aussi bien pour un nettoyage et une désinfection
mécaniques que manuels.
Traitement préalable
Éliminer toute contamination immédiatement après
utilisation ou dans les 2 heures suivantes. Pour ce faire,
rincer minutieusement le conducteur de lumière sous
l‘eau courante (pendant au moins 10 secondes). Il est
également possible d‘utiliser une solution de désinfection
sans aldéhyde adaptée pour éviter la fixation de sang.
Pour éliminer la contamination manuellement, utiliser
un pinceau ou un chiffon doux. Un composite
partiellement polymérisé peut être éliminé avec de
l‘alcool et à l‘aide d‘une spatule en plastique. Ne pas
utiliser d‘objets coupants ou pointus, ceux-ci
pour raient endommager la surface.
Nettoyage et désinfection
Pour le nettoyage, immerger l’embout lumineux dans une
solution de nettoyage de manière à ce qu’il soit
suffisamment recouvert par le liquide (des ultrasons ou un
brossage délicat avec un pinceau doux peuvent renforcer
le nettoyage). Un agent de nettoyage neutre-enzymatique
est recommandé. Lors du nettoyage et de la désinfection,
vérifier que les agents utilisés sont exempts :
d‘acides organique, minéral et oxydant (la valeur pH
minimum admissible est 5,5)
de solutions alcalines (la valeur pH maximum
admissible est 8,5)
d‘agents oxydants (par ex. peroxydes d‘hydrogène).
Enlever ensuite l’embout lumineux de la solution et le
rincer minutieusement à l’eau courante (pendant au
moins 10 secondes). Un nettoyage dans un stérilisateur
thermique est une alternative efficace.
Stérilisation
Un nettoyage et une désinfection intensifs sont les
préalables indispensables pour assurer une stérilisation
efficace. Utiliser pour cela la stérilisation en autoclave. Le
temps de stérilisation (temps d’exposition à température
de stérilisation) est de 4 minutes à 134 °C; la pression
doit être de 2 bar. Sécher l’embout lumineux stérilisé en
utilisant soit le programme de séchage scial de l’auto-
clave, soit de l’air chaud. L’embout lumineux a été tes
pour résister jusqu’à 200 cycles de stérilisation.
Vérifier ensuite que le conducteur de lumière n‘ait subi
aucun dommage. Le tenir à contre-jour. Si certains
segments deviennent noirs, c‘est que les fibres de verre
sont cassées. Dans ce cas, remplacer le conducteur de
lumière par un neuf.
Contacts de la batterie
Pour garantir une conductivité fiable et constante, éviter
toute contamination des contacts de la batterie (par
exemple par des résidus de composite). Pour cela, nettoyer
régulièrement les contacts lors de la désinfection
habituelle (après chaque patient).
Élimination des pièces usagées
Ne pas jeter la lampe à photopolymériser avec
les ordures ménagères. Le recyclage doit se
faire conformément à la réglementation et
aux dispositions nationales. Ne pas incinérer
les batteries.
28
Que faire en cas de …?
Symbole Causes Rectification de l‘erreur
Défaut d‘électronique dans la pièce à
main ou la batterie
L‘appareil a surchauffé ou a subi une
température trop basse
Défaut électronique dans le batterie
Batterie vide
Les contacts sont contaminés
Batterie en mode sécurité
(circuit de protection intégré)
Le Power pack n‘est pas connecté ou
est défectueux.
Retirer et réinsérer la batterie. Si l‘erreur persiste, remplacer
la batterie par le power pack (Click & Cure).
Si l‘erreur persiste, veuillez contacter votre distributeur ou le
service après-vente local.
Laisser l‘appareil refroidir (ou le laisser remonter à tempéra-
ture ambiante s‘il est trop froid) et essayer de nouveau
après un certain délai.
Si l‘erreur persiste, veuillez contacter votre distributeur ou le
service après-vente local.
Retirer et réinsérer la batterie. Si l‘erreur persiste, placer la
pièce à main sur la base de chargement.
Si malgré cela, l‘erreur persiste toujours, remplacer la
batterie par le power pack (Click & Cure). Veuillez contacter
votre distributeur ou le service après-vente local.
Positionner la pièce à main correctement dans la base de
chargement. „Charging Battery“ s‘affiche brièvement avant
le retour en mode stand-by de l‘écran. Si la batterie n‘est
pas mise en charge, nettoyer les contacts. Si l‘erreur
persiste, veuillez contacter votre distributeur ou le service
après-vente local.
Vérifier que la pièce à main est correctement en place dans
la base de chargement. Nettoyer les contacts. Information :
„Charging Battery“ s‘affiche brièvement avant le retour en
mode stand-by de l‘écran.
Si nettoyer les contacts n‘améliore pas les résultats, retirer
la batterie de la pièce à main. Séparer la batterie de la
pièce à main et la mettre en charge dans la base de
chargement.
Si l‘erreur persiste, veuillez contacter votre distributeur ou le
service après-vente local.
Vérifier que le power pack est correctement positionné dans
la base de chargement (vérifier également la position des
contacts), et qu‘il est raccordé à la prise de courant par le
cordon électrique (le témoin lumineux sur le power pack
s‘allume en vert s‘il fonctionne correctement).
Si l‘erreur persiste, veuillez contacter votre distributeur ou le
service après-vente local.
Pas
d‘affichage
pendant le
chargement
La LED
de la base
de charge-
ment ne
s‘allume
pas
29
Garantie /
Procédure en cas de réparation
Spécifications du produit
La lampe Bluephase N est garantie 3 ans à partir de la
date d‘achat (batterie 1 an).
Des pannes dues à un matériel défectueux ou à des vices
de fabrication sont réparées gratuitement pendant la
durée de la garantie. La garantie ne couvre aucun
dommage, matériel ou non, autre que ceux mentionnés.
L‘appareil doit être utilisé exclusivement pour les indica-
tions définies. Tout autre usage est contre-indiqué. Le
fabricant rejette toute responsabilité quant à un mauvais
usage de l‘appareil. Aucune demande de garantie ne
peut être acceptée dans ce genre de cas. Ceci est
particulièrement valable en cas de :
dommage lié à un usage inadapté, spécialement dans
le cas de batteries mal conservées (voir Batterie);
dommage de composant lié à l‘usure dans les
conditions d‘utilisation normale (voir Batterie);
dommage lié à des facteurs externes, exemple un
choc, une chute sur le sol;
dommage lié à des réglages ou à une installation
incorrects;
dommage survenant lors du branchement de la lampe
sur une source d‘énergie dont le voltage et la
fréquence ne sont pas conformes à ceux inscrits sur
le socle de la lampe.
dommage lié à des réparations inadaptées ou à des
modifications réalisées par des réparateurs non
certifiés.
En cas de réclamation sous garantie, la lampe complète
(pièce à main, base de chargement, cordon électrique, et
le power pack) doivent être retournés en port dû au
distributeur ou directement à Ivoclar Vivadent, avec la
facture d‘achat. Utiliser le conditionnement d‘origine et
le carton correspondant pour le transport.
Les réparations doivent être réalisées uniquement par un
service certifié par le service après-vente Ivoclar
Vivadent. En cas de défaut ne pouvant être rectifié,
veuillez contacter votre distributeur ou le service après-
vente local (voir les adresses au dos). Une description
claire du défaut ou des conditions de sa survenue
pourront faciliter l‘identification du problème. Veuillez
joindre ce descriptif quand vous retournez l‘appareil.
Présentation
1 Base de chargement,1 Cordon électrique, 1 Pièce à main,
1 Batterie (Li-Po), 1 Conducteur de lumière 10 mm, 3 Embouts de
protection oculaire, 1 Mode d‘emploi
Accessoires
Les accessoires suivants sont disponibles pour la Bluephase N :
REF Description
656196 Conducteur de lumière 10 mm, noir (N)
608538 Conducteur de lumière Pin-point 6>2 mm, noir
551756 Embouts de protection oculaire
592496 Disque de protection oculaire
659956 Batterie Bluephase N
659681 Pièce à main Bluephase N
607922 Radiomètre Bluephase Meter
608554 Housses de protection Bluephase
Données techniques
Voltage base de chargement 5 VDC
Voltage pièce à main 3.7 VDC avec la batterie
5 VDC avec le power pack
Power pack 100–240 VAC / 50-60 Hz /
max. 0.4 A
Fabricant: Friwo
Puissance 5 VDC
TYPE 15.2630
Conditions de fonctionnement
Température +10 °C à +35 °C
Humidité relative 30% à 75%
Pression ambiante 700 hPa à 1060 hPa
Dimensions de la base de chargement L = 205 x W = 150 x H = 85 (mm)
Poids de la base de chargement 250 g
Temps de chargement env. 2 h (avec une batterie vide)
Source électrique de la pièce à main Batterie Li-Po
Temps de batterie maximum env. 60 min.
(avec une batterie neuve et
totalement chargée)
Source lumineuse Polywave
®
LED
Spectre 385 – 515 nm
Intensité lumineuse max. 1'200 mW/cm
2
± 10 %
Fonctionne 5 min. on / 6 min. off
(par intermittence)
Dimensions de la pièce à main L = 260 x W = 42 x H = 120 (mm)
Poids de la pièce à main (batterie incluse) 225 g
Transport et conditions de stockage :
Température -20 °C à +60 °C
Humidité relative 10% à 75%
Pression ambiante 500 hPa à 1060 hPa
Bluephase N, doit être rangée en hauteur et dans une pièce fermée.
Protéger l‘appareil des chocs importants.
Batterie
Ne pas stocker à des températures supérieures à 40 °C (ou 60 °C
pour une courte période). Température de stockage recommandée
15 – 30 °C.
Toujours stocker une batterie chargée
et jamais pendant plus 6 mois.
30
Estimado cliente,
Una óptima polimerización es un requisito
importante para todos los materiales fotopolimeri-
zables con el fin de realizar duraderas restauracio-
nes de alta calidad. La lámpara de polimerización
seleccionada también juega un papel decisivo a
este respecto. Por ello, nos gustaría agradecerle el
que usted haya comprado Bluephase N. Bluephase
N es un dispositivo médico de alta calidad que ha
sido diseñado de acuerdo con los últimos estánda-
res de la ciencia y tecnología en conformidad con
importantes estándares industriales.
Estas instrucciones de uso le ayudarán a poner en
marcha la lámpara de forma segura, aprovechar
todas sus posibilidades y asegurar una larga vida
útil.
Si tuviera usted alguna pregunta adicional, no
dude en contactarnos (consulte las direcciones en
la página del reverso).
Su equipo de Ivoclar Vivadent
31
Español
Producto 32
– Despiece
Indicadores en la base de carga
Indicadores en la pieza de mano
Seguridad 34
Uso previsto
– Indicaciones
Aclaración de símbolos
Avisos de seguridad
Puesta en marcha 36
Base de carga
Pieza de mano
– Batería
Funcionamiento con cable Click&Cure
Funcionamiento 39
Selección del programa de polimerización y tiempo de polimerización
Función de memoria de polimerización
– Inicio
Señales acústicas
Intensidad lumínica
Medición de la intensidad lumínica
Mantenimiento y limpieza 41
– Carcasa
Conductos de luz
Contactos de la batería
– Desecho
¿Qué sucede si …? 42
Garantía 43
Procedimiento en caso de reparación 43
Especificaciones del producto 43
Índice
32
Despiece
1 Pieza de mano
2 Interruptor de puesta en marcha
3 Pantalla
4 Tecla de selección de programa
5 Tecla de selección de tiempo
6 Tecla de volumen
7 Indicador de potencia
8 Base de carga
9 Cable eléctrico
10 Unidad de alimentación
11 Batería
12 Conducto de luz 10 mm
13 Cono antideslumbrante
10
11
311213
2
9
7 8
5 6 4
33
Indicadores en la base de carga
Indicadores en la pieza de mano
Indicador se ilumina azul – la base de carga está
conectada a la red
Programa de polimerización y estado de funcionamiento
respectivamente
Indicaciones seleccionadas y tiempo de polimerización
respectivamente
Estado de carga de la batería
34
Uso previsto
Bluephase N es una lámpara de polimerización LED que
produce luz azul de alta energía. Se utiliza para la poli-
merización de los materiales dentales fotopolimerizables
directamente en el sillón dental. El uso previsto también
incluye la observación de estas instrucciones de uso.
Indicaciones
Con su espectro de banda ancha “multiondas”,
Bluephase N está indicada para la polimerización de
todos los materiales dentales fotopolimerizables en la
franja de longitud de onda de 385–515 nm. Estos
materiales incluyen materiales de restauración, agentes
adhesivos, bases, liners, selladores de fisuras, provisiona-
les, así como materiales de cementación para brackets y
trabajos de laboratorio tales como inlays de cerámica.
Aclaración de símbolos
Uso no permitido
Símbolos en la lámpara
Doble aislamiento (El dispositivo cumple con
la Clase de Seguridad II)
Protección frente al choque eléctrico
(aparato tipo BF)
Observar las instrucciones de uso
Observar las instrucciones de uso
Precaución
El aparato no se puede eliminar con la
basura doméstica normal. Por favor, visite
la página Web de Ivoclar Vivadent de su
país, para información más detallada sobre
la manera de desechar las lámparas.
Reciclable
Notas de seguridad
Bluephase N es un dispositivo electrónico y un producto
sanitario sujeto a las directivas IEC 60601-1 (EN 60601-
1) y EMC, además de las Directivas de Disposi tivos
Médicos 93/42/EEC. El aparato cumple con las normas
vigentes de la UE y está clasificado como producto LED
de Clase 2.
El aparato has sido enviado desde el fabricante en condi-
ciones técnicas y de seguridad optimas. Para mantener
dichas condiciones y asegurar un funcionamiento seguro,
se deben tener en cuenta las notas y regulaciones de
estas instrucciones de uso. Para evitar daños al equipo y
riesgos para los pacientes, los usuarios y terceras perso-
nas, deben tener en cuenta las siguientes instrucciones
de seguridad.
Los materiales que se polimerizan fuera de la
franja de longitud de onda de 385–515 nm
(hasta la fecha no se conocen). Si usted no
está seguro acerca de cierto productos, por
favor pregunte al fabricante del correspon-
diente material.
No cargar ni utilizar el aparato cerca de
sustancias inflamables o combustibles.
Los dispositivos de comunicación portátiles o
móviles de alta frecuencia pueden interferir
con el equipamiento médico. No está permi-
tido el uso de teléfonos móviles durante el
funcionamiento de la lámpara.
Seguridad
35
Uso y Responsabilidad
Bluephase N solo se puede utilizar para el uso previsto
y cualquier otro uso está contraindicado. No se puede
aceptar responsabilidad por daños que resulten de un
uso inadecuado o de la no observancia de las instruc-
ciones de uso.
El usuario tiene la responsabilidad de revisar
Bluephase N en cuanto a su uso e idoneidad para los
usos previstos. Ello es especialmente importante si se
utilizan al mismo tiempo otros equipos en inmediata
proximidad de Bluephase N.
Utilice únicamente piezas de repuesto y accesorios
originales de Ivoclar Vivadent (ver accesorios). El fabri-
cante no acepta responsabilidad alguna por los daños
que se puedan producir si se utilizan piezas de repues-
tos y accesorios de otros fabricantes.
El conducto de luz es una pieza aplicada y puede
alcanzar una temperatura máxima de 45 °C (113 °F)
en el punto de contacto con la pieza de mano durante
su funcionamiento.
Voltaje de funcionamiento
Antes de la conexión, asegúrese de que:
a) el voltaje indicado en la placa de características
cumpla con el suministro energético local.
b) la unidad haya alcanzado la temperatura ambiente.
Si se utilizan por separado la batería y la unidad alimen-
tadora, e.g. durante la puesta en marcha o el funciona-
miento con cable Click&Cure, se debe evitar el contacto
con los pacientes o terceras partes. No toque los contac-
tos expuestos de la batería o de la clavija de conexión
(unidad alimentadora).
Supuesto de seguridad perjudicial
Si se sospecha que el funcionamiento ya no puede ser
seguro, se debe desenchufar de la red eléctrica para
evitar un funcionamiento accidental. Esto se podría dar,
por ejemplo, si el aparato está visiblemente dañado o ya
no funciona correctamente. La interrupción completa de
electricidad solo queda asegurada desenchufando el
mismo de la red.
Protección ocular
Se debe evitar la exposición directa o indirecta de los
ojos. La exposición prolongada de las lámparas es desa-
gradable para la vista y puede producir daños. Por ello,
se recomienda el uso de los conos antideslumbrantes
incluidos. Los individuos que por lo general sean sensi-
bles a la luz, los que tomen medicamentos fotosensibili-
zantes o los que se hayan sometido a cirugía ocular, o
las personas que trabajen con el aparato o en su proxi-
midad durante largos períodos de tiempo no deberán
exponerse a la luz de este dispositivo y llevar gafas pro-
tectoras (naranjas) que absorben la luz por debajo de los
515 nm.
Batería
Protección: Utilice únicamente piezas de repuestos origi-
nales, especialmente baterías de Ivoclar Vivadent y bases
de carga. No cortocircuite la batería. No lo almacene a
temperaturas por encima de los 40º C/104º F (o 60º C/
140º F durante un corto período de tiempo). Siempre
almacene las baterías cargadas. El período de almacena-
miento no debe exceder los 6 meses. Puede explotar si
se elimina en fuego.
Por favor, tenga en cuenta que la bateria de polímero de
litio puede reaccionar con explosión, fuego, y producir
humo en caso de una manipulación inapropiada y daños
mecánicos. Baterias de polímero de litio dañadas, no
deben volver a utilizarse
Los electrolitos y gases de electrolitos liberados durante
explosión, fuego y desarrollo de humo son tóxicos y
corrosivos. En caso de contacto con los ojos y piel lavar
inmediatamente con abundante cantidad de agua. Evitar
la inhalación de humos. Acudir a un especilaista
inmeditamente en caso de indisposición.
Desarrollo térmico
Al igual que sucede con todas las lámpara de alto
rendimiento, la alta intensidad lumínica provoca cierto
desarrollo térmico. La prolongada exposición de áreas
próximas a pulpa y tejidos blandos puede provocar
daños irreversibles o reversibles. Por ello, se deben tener
en cuenta los tiempos de polimerización recomendados y
se deben evitar tiempos de polimerización ininterrumpi-
dos de más de 20 segundos sobre la misma superficie
dental, así como el contacto directo con la gingiva, mem-
brana mucosa dental o piel. Si fuera necesario, polimerice
restauraciones indirectas en intervalos intermitentes de
20 segundos cada vez o utilice refrigeración externa con
pistola de aire.
36
Revise el suministro en cuanto a su integridad y cual-
quier posible daño de transporte (ver forma de suminis-
tro). Si faltaran componentes o estuvieran dañados,
contacte inmediatamente con su distribuidor o su servicio
técnico local.
Base de carga
Antes de conectar el dispositivo, asegúrese de que el vol-
taje indicado en la placa de características cumpla con el
suministro energético local. La placa de características
está situada en la parte inferior de la base de carga.
Coloque la estación de carga encima de una mesa ade-
cuada y plana. Retire el folio protector de la clavija de
conexión de la unidad de alimentación. Conecte la clavija
de conexión de la unidad de alimentación en el enchufe
en la parte inferior de la base de carga. Inclínela ligera-
mente y aplique una ligera presión hasta que escuche
como encaja correctamente.
Conecte el cable eléctrico al suministro energético y a la
unidad de alimentación. El indicador de potencia ON en
el lado izquierdo de la carcasa se ilumina en azul (ver
indicadores en la base de carga).
Pieza de mano
Antes de insertar el conducto de luz, retire el folio pro-
tector en la correspondiente apertura de la pieza de
mano.
Gire ligeramente el conducto de luz mientras se inserta
en la pieza de mano.
Seguidamente, coloque el cono antideslumbrante en el
conducto de luz.
Puesta en marcha
37
Batería
Recomendamos cargar completamente la batería antes
de su primer uso. Si la batería está completamente
cargada, presenta una capacidad de polimerización de
aproximadamente 60 minutos.
Deslice la batería recta en la pieza de mano hasta que
escuche que encaja perfectamente.
Colocar la pieza de mano con suavidad en el apoyo
correspondiente de la base de carga. Si se utilizó la
funda, por favor retírela antes de cargar la bateria.
Si fuera posible, utilice la lámpara siempre con una bate-
ría completamente cargada, ya que ello prolongará la
vida útil. Por ello se recomienda colocar la pieza de
mano en la base de carga después de cada paciente. El
tiempo de carga para baterias vacias es de 2 horas.
Estado de carga de la batería
Con la pieza de mano conectada, se muestra el actual
estado de carga en la pantalla como sigue:
Batería completamente cargada
(capacidad de polimerización de aproxima-
damente 60 minutos)
Batería medio llena
Reserva (La última barra en el indicador de
batería está rojo en modo de reserva. La
batería se debe cargar lo antes posible.)
Cuando la bateria se está cargando, aparece
brevemente en el display „Charging Battery“
antes de que la pantalla regrese al modo
stand-by
Batería vacía
Si la batería está completamente vacía, la
pieza de mano se desconecta automática-
mente a modo stand-by. La lámpara ya no
se puede conectar y el programa y tiempo
de polimerización tampoco se pueden
ajustar. Sin embargo, la pieza de mano se
puede utilizar en modo de funcionamiento
con cable Click&Cure.
Ya que la batería es un consumible, ésta se debe
re emplazar después de su típico ciclo útil de 2½ años.
Ver etiqueta en la batería para comprobar la edad de la
misma.
38
Funcionamiento con cable Click&Cure
En cualquier momento se puede utilizar Bluephase N con
funcionamiento con cable, pero especialmente si la
batería está completamente vacía.
Para dicho fin, retire la batería de la pieza de mano y
seguidamente retire la unidad alimentadora de la parte
inferior de la base de carga. No tire del cable eléctrico.
Inserte la clavija de conexión en la pieza de mano hasta
que oiga y sienta que ha encajado en su posición.
Durante la operación con cable, la base de carga no puede
cargar la batería, ya que no está conectada a una fuente
de energía.
El aparato está completamente desconectado del sumi-
nistro de alimentación sólo cuando el cable eléctrico ha
sido desconectado del enchufe.
Desinfecte o autoclave las superficies contaminadas de la
lámpara de polimerización, así como los conductos de luz
y conos antirreflectantes antes de cada uso. Además,
asegúrese de que la intensidad lumínica estipulada per-
mita una adecuada polimerización. Para ello, revise el
conducto de luz en cuanto a contaminación o daños, así
como la intensidad lumínica en intervalos regulares (e.g.
con el radiómetro de Ivoclar Vivadent “Bluephase
Meter”).
Selección del programa y tiempo
de polimerización
El programa de polimerización y el tiempo de polimeriza-
ción se pueden ajustar individualmente. Bluephase N
está equipada con los tres siguientes programas de poli-
merización para las diferentes indicaciones. El programa
de polimerización deseado se selecciona con las teclas de
selección de programa. La pantalla cambia como
corresponde (ver indicadores en la pieza de mano).
Después de la primera puesta en marcha, se han prede-
terminado los siguientes ajustes:
HIGH (Potencia alta): 10 segundos
LOW (Potencia baja): 10 segundos
SOFT (Inicio suave): 15 segundos.
El tiempo de polimerización deseado se elige utilizando
las teclas de selección de tiempo. Los usuarios pueden
elegir entre 5, 10, 15, 20 y 30 segundos.
Tenga en cuenta las instrucciones de uso del material
aplicado a la hora de seleccionar el tiempo de polimeri-
zación. Las recomendaciones de polimerización para los
materiales de composites se aplican para todos los colo-
res y, si no se indica lo contrario en las instrucciones de
uso del material pertinente, a un grosor de capa máximo
de 2 mm. Por lo general, estás recomendaciones se apli-
can a situaciones en las que el extremo del conducto de
luz se coloca directamente sobre el material que se va a
polimerizar. Si se aumenta la distancia entre la fuente de
luz y el material, el tiempo de polimerización se ampliará
de forma correspondiente. Por ejemplo, si la distancia al
material es de 9 mm, la salida efectiva de luz se reduce
en aproximadamente un 50%. En este caso, el tiempo de
polimerización recomendado se debe duplicar.
Funcionamiento
39
Programas de polimerización
1) Aplicar hasta un máximo de 2 mm si en las instrucciones de uso no recomiendan ninguna otra
acción (como es el caso, ej. de colores de dentina)
2) Aplicar hasta un máximo de 4 mm si en las instrucciones de uso no recomiendan ninguna otra
acción (como es el caso, ej. de colores de dentina)
3) Aplicar a un grosor de capa máximo de 3mm
4) Aplicar fotopolimerización (solo en pasta base)
5) Aplicar a polimerización dual
Programa HIGH POWER
Consistente alta intensidad lumínica
para la polimerización de materiales
de restauración y cementación para
restauraciones directas e indirectas.
Programa SOFT START
Aumento gradual de la intensidad
lumínica con reducida tensión de
polimerización y reducido desarrollo
térmico para la polimerización de
materiales de restauración.
Programa LOW POWER
Reducida intensidad lumínica con
reducido desarrollo térmico para la
polimerización de adhesivos, liners, y
materiales de restauración en áreas
próximas a pulpa al restaurar cavi-
dades Clase V.
650 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0 5
650 mW/cm
2
Materiales de restauración
Composite
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
Todos los composites convencionales
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Compómeros
3)
Compoglass F / Compoglass Flow
Restauraciones indirectas /
Materiales de cementación
Variolink II Base
4)
/ Variolink Veneer / Variolink N Base
4)
/ Variolink N Clear Veneer / Variolink Ultra
5)
/
Dual Cement
5)
/ Variolink II
5)
Otros
Helioseal / Helioseal F / Helioseal Clear
Monopaque
MultiCore Flow / Multicore HB
Heliosit Orthodontic
Telio Add-On Flow
Telio Stains
IPS Empress Direct Color
IPS Empress Direct Opaque
Tiempo de polimerización
10 seg.
15 seg.
10 seg.
20 seg.
por mm cerámica:
10 seg. por segmento
10 seg.
20 seg.
20 seg.
10 seg.
15 seg.
10 seg.
10 seg.
20 seg.
Adhesivos
AdheSE / AdheSE One F
ExciTE / ExciTE DSC
Tetric N-Bond / Tetric N-Bond Self-Etch
Heliobond (Syntac)
Materiales provisionales
Telio CS Inlay/Onlay
Systemp.inlay/onlay
Fermit / Fermit N
Telio CS Link / Systemp.link
Otros
Heliosit Orthodontic
Vivaglass Liner
Tiempo de polimerización
10 seg.
10 seg.
10 seg.
10 seg.
20 seg. por segmento
20 seg.
20 seg.
Materiales de restauración
Composite
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
Todos los composites convencionales
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Compómeros
3)
Compoglass F / Compoglass Flow
Tiempo de polimerización
15 seg.
20 seg.
15 seg.
20 seg.
Intensidad de luz
Tiempo de exposición para composites
Tiempo de exposición para Tetric EvoCeram / IPS Empress Direct
1.200 mW/cm
2
15 seg.
10 seg.
40
Función de memoria de polimerización
Los últimos ajustes utilizados, junto con la combinación
del programa y tiempo de polimerización, se almacenan
automáticamente.
Inicio
La lámpara se conecta con la tecla “Start”. Una vez
transcurrido el tiempo de polimerización seleccionado, el
programa de polimerización finaliza automáticamente.
Si se desea, la lámpara se puede desconectar antes de
finalizar el tiempo de polimerización ajustado, volviendo
a pulsar la tecla “Start”. El ventilador se activa al mismo
tiempo que la lámpara. Una vez transcurrido el tiempo
de polimerización, el ventilador sigue funcionando duran-
te cierto tiempo para refrigerar el aparato. La batería no
se debe retirar mientras el ventilador siga funcionando.
Señales acústicas
Las señales acústicas son audibles para las siguientes
funciones:
Inicio (Parada)
Cada 10 segundos
Cambio de programa
Cambio de tiempo de polimerización
Conexión de la batería a la base de carga
Inserción de la batería
Mensaje de error
Si se desea, se puede ajustar el volumen de las señales
acústicas. Para ello, pulse la tecla de volumen azul mien-
tras que la lámpara está apagada y reduzca el volumen o
desconecte las señales acústicas con la tecla de progra-
ma izquierda. Si desea conectar de nuevo las señales
acústicas o incrementar el volumen, pulse la tecla azul
de volumen y entonces, pulse la tecla derecha de selec-
ción de programa.
Intensidad lumínica
La intensidad lumínica se mantiene a un nivel consisten-
te durante el funcionamiento. Si se utiliza el conducto de
luz adjunto de 10 mm, la intensidad lumínica debe
calibrarse a 1200 mW/cm
2
± 10%.
La utilización de otros conductos de luz diferentes de los
suministrados tiene una influencia directa sobre la inten-
sidad lumínica emitida.
En conductos de luz de paredes paralelas (10 mm), el
diámetro es igual en ambos extremos. En conductos de
luz que focalizan (conducto de luz 10>8 mm, conducto
Pin Point 6>2 mm), el diámetro del extremo interior es
mayor que el de la ventana de emisión de luz. Así, el haz
de luz azul incidente se reduce a una superficie más
pequeña, lo que aumenta la intensidad lumínica emitida.
Los conductos de luz Pin-Point están indicados para la
polimerización de zonas reducidas, tales como la fijación
de carillas antes de la eliminación del sobrante de mate-
rial. Para una polimerización correcta completa, es nece-
sario cambiar el conducto de luz.
Medición de la intensidad lumínica
La intensidad lumínica de Bluephase N y el conducto de
luz adjunto de 10 mm se pueden revisar con Bluephase
Meter.
Si el valor medido no se corresponde a la intensidad
lumínica esperada, siga los siguientes pasos:
Revise el programa de polimerización seleccionado
Limpie la posible suciedad del fotosensor (radiómetro)
Retire el conducto de luz y limpie la boquilla de emi-
sión de luz de la pieza de mano con una torunda de
algodón humedecida en alcohol.
Limpie la posible suciedad del conducto de luz (consultar
mantenimiento y limpieza)
Reemplace el conducto de luz dañado con uno nuevo.
Si estas medidas no mejoran los resultados, por favor
póngase en contacto con su distribuidor o con su servicio
técnico local.
41
Por razones de higiene, se recomienda utilizar un protector
plástico desechable para cada paciente. Asegúrese de
colocar el protector bien ajustado al conducto de luz. Si no
utiliza protectores plásticos desechables, des infecte las
superficies contaminadas del dispositivo y los conos
antideslumbrantes (FD 366/Dürr Dental, Incidin Liquid/
Ecolab) y esterilice el conducto de luz antes de cada uso.
Durante la limpieza del dispositivo, asegúrese de que
ningún líquido ni sustancia externa penetra en la pieza de
mano, la base de carga y especialmente al transformador
(riesgo de descarga eléctrica). Desconecte la base de carga
de la fuente de alimentación para limpiarla.
Cubierta
Limpie la pieza de mano y el soporte de la pieza de mano
con una solución desinfectante común, libre de aldehídos.
No limpie con soluciones desinfectantes alta mente agresi-
vas (por ejemplo, soluciones a base de aceite de naranja
o con un contenido de etanol superior al 40 %), solventes
(por ejemplo, acetona) o instrumentos punzantes que
podrían dañar o rayar el plástico. Limpie las piezas plásticas
con una solución jabonosa.
Conducto de luz
Antes de limpiar y/o desinfectar el conducto de luz, es
ne cesario realizar un tratamiento previo. Esto se aplica
tanto a la limpieza como a la desinfección, ya sea
automática o manual.
Tratamiento previo
Quite la suciedad visible inmediatamente después del
uso o dentro de las 2 horas posteriores. Para ello,
enjuague abundantemente el conducto de luz con
agua corriente (durante, por lo menos, 10 segundos).
También puede utilizar una solución desinfectante
libre de aldehídos para evitar que se adhiera la sangre.
Para eliminar la suciedad en forma manual, utilice un
cepillo de cerdas blandas o un paño suave. Los com-
puestos parcialmente polimerizados pueden eliminarse
con alcohol y una espátula de plástico, si es necesario.
No utilice objetos afilados o puntiagudos, ya que
podrían rayar la superficie.
Limpieza y desinfección
Para limpiar, sumerja el conducto de luz en una solución
de limpieza y asegúrese de cubrirlo por completo con el
líquido (el ultrasonido o un ligero cepillado con un cepillo
de cerdas blandas podría mejorar el efecto). Se recomienda
utilizar un agente de limpieza enzimático neutro. Al limpiar
y desinfectar, asegúrese de que los agentes utilizados no
contengan:
ácidos orgánicos, minerales y oxidantes (el valor de pH
mínimo admitido es de 5,5)
solución alcalina (el valor de pH máximo admitido es
de 8,5)
agente oxidante (por ejemplo, peróxido de hidrógeno)
Luego, retire el conducto de luz de la solución y enjuáguelo
con abundante agua corriente (durante por lo menos
10 segundos). La limpieza con un desinfectante térmico es
una alternativa eficaz.
Esterilización
Durante la fase de limpieza y desinfección, es impres-
cindible asegurarse de que la posterior esterilización sea
eficaz. Para ello, utilice únicamente el método de esteri-
lización por autoclave. El tiempo de esterilización (tiempo
de exposición a temperatura de esterilizado) es de
4 minutos, a 134 °C (273 °F); la presión debe ser de
2 bares (29 psi). Seque el conducto de luz esterilizado
con el programa de secado especial del autoclave a vapor
o aire caliente. El conducto de luz ha sido probado para
hasta 200 ciclos de esterilización. Luego, controle que el
conducto de luz no esté dañado. Sujételo a contraluz. Si
los segmentos individuales se ven negros, las fibras de
vidrio están quebradas. En ese caso, cambie el conducto
de luz por uno nuevo.
Contactos de la batería
Para asegurar una conductividad confiable en todo
momento, mantenga los contactos de la batería libres de
posibles agentes contaminantes (por ejemplo, residuos
de composite). Para ello, limpie los contactos utilizados
regularmente durante la desinfección de limpieza normal
(después de cada paciente).
Eliminación
La lámpara de polimerización no debe
desecharse junto con los residuos urbanos.
Elimine las baterías fuera de uso y las lámpa-
ras de poli merización de acuerdo con las nor-
mas legales correspondientes de su país. Las
baterías no deben incinerarse.
Mantenimiento y limpieza
42
¿Qué sucede si …?
Símbolo Causas Subsanar el error
Defecto electrónico en la pieza de
mano o batería
El aparato está sobrecalentado o
sobre enfriado
Defecto electrónico en la batería
Batería vacía
Los contactos de carga están
contaminados
Bateria en modo seguridad
(circuito de protección integrado)
La unidad alimentadora no está
conectada o está defectuosa
Retire y vuelva a colocar la batería. Si el error persiste,
reemplace la batería con la unidad alimentadora (Click & Cure).
Si el error persiste, por favor contacte con su distribuidor o su
servicio técnico local.
Deje que el aparato se refrigere (o que alcance temperatura
ambiente si estuviera sobre enfriado) y vuelva a intentarlo des-
pués de un tiempo. Si el error persiste, por favor contacte con
su distribuidor o servicio técnico local.
Retire y vuelva a colocar la batería. Si el error persiste, coloque
el aparato en la base de carga.
Si a pesar de ello, el error persiste, reemplace la batería con la
unidad alimentadora (Click & Cure). Por favor, contacte con su
proveedor o servicio técnico local.
Posición de la pieza de mano en la base de carga.
Aparece brevemente en el display „Charging Battery“ antes de
que la pantalla regrese al modo stand-by. Si la batería no se
carga, se deben limpiar los contactos. Si el error persiste, por
favor contacte con su distribuidor o servicio técnico local.
Revise si la pieza de mano está correctamente colocada en la
base de carga. Limpie los contactos de la batería. Información:
Aparece „Charging Battery“ brevemente en el display antes
que la pantalla regrese al modo stand-by.
Si al limpiar los contactos de la batería no mejoran los resulta-
dos, retire la batería del aparato. Cargue la batería indepen-
dientemente de la pieza de mano en la base de carga durante
aproximadamente 10 minutos. Si el error persiste, por favor
contacte con su distribuidor o su servicio técnico local.
Revise si el cargador está correctamente colocado en la base de
carga (revise también los contactos de carga) o si el cargador
está conectado con el cable a la corriente eléctrica (si funciona
correctamente, los indicadores del cargador están en verdes).
Si el error persiste, por favor contacte con su distribuidor o su
servicio técnico local.
No Display
durante la
carga
No se
ilumina el
LED de la
base de
carga
43
Garantía /
Procedimiento en caso de reparación
Especificaciones del
producto
El período de garantía de Bluephase N es de 3 años
desde la fecha de compra (1 año para la batería).
Las averías que se produzcan por material defectuoso o
fallos de fabricación se reparan gratis durante el período
de garantía. La garantía no otorga el derecho a recuperar
cualquier daño de material o no material distinto de los
mencionados. El aparato solo se debe utilizar para los
fines previstos. Cualquier otro uso está contraindicado. El
fabricante no acepta responsabilidad derivada de un uso
inadecuado y, en tales casos no se puede aceptar las
reclamaciones de la garantía. Ello es particularmente
válido para:
daños producidos por una manipulación inadecuada,
especialmente de baterías almacenadas incorrecta-
mente (ver datos técnicos: Condiciones de transporte y
almacenamiento).
daños de los componentes resultado de utilización
por debajo de los estandares de las condiciones de
funcionamiento (ej baterias)
daños producidos por influencias externas,
e.g. explosiones, caídas al suelo
daños producidos de una incorrecta puesta en
marcha o instalación
daños producidos por conectar el dispositivo a una
red eléctrica cuyo voltaje y frecuencia no cumplan
con las indicadas en la placa de características
daños producidos por reparaciones inadecuadas o
modificaciones que no hayan sido realizadas por
servicios técnicos autorizados
En caso de reclamación bajo garantía, el aparato
completo (pieza de mano, base de carga, cordón
eléctrico y unidad alimentadora) se debe enviar portes
pagados a su distribuidor o directamente a Ivoclar
Vivadent, junto con el justificante de compra. Utilice el
envase original con los correspondientes cartones de
embalaje para su transporte.
Los trabajos de reparación sólo se pueden realizar en un
servicio técnico autorizado por Ivoclar Vivadent. En el
caso de una avería que no se pueda corregir, por favor
contacte con su distribuidor o servicio técnico local (con-
sultar direcciones en el reverso). Una clara descripción de
la avería o las condiciones bajo las cuales se ha produci-
do la misma, facilitará el diagnóstico del problema. Por
favor, incluya dicha descripción cuando envíe el aparato.
Forma de suministro
1 Base de carga, 1 Cable eléctrico, 1 Unidad de alimentación, 1 Pieza
de mano, 1 Batería (Li-Po), 1 Conducto de luz de 10 mm, 3 conos anti-
deslumbrantes, 1 juego de instrucciones de uso
Accesorios
Se dispone de los siguientes accesorios para Bluephase N
REF Descripción
656196 Conducto de luz 10 mm, negro (N)
608538 Conducto de luz Pin Point 6>2 mm, negro
551756 Conos antideslumbrantes
592496 Pantalla antideslumbrante
659956 Bluephase N batería
659681 Bluephase N pieza de mano
607922 Bluephase Meter
608554 Bluephase fundas
Datos técnicos
Voltaje de funcionamiento base de carga 5 VDC
Voltaje de funcionamiento pieza de mano 3.7 VDC con batería
5 VDC con unidad alimentadora
Unidad alimentadora 100-240 VAC / 50-60 Hz / máx. 0.4 A
Fabricante Friwo
Salida 5 VDC
TIPO 15.2630
Condiciones de funcionamiento
Temperatura + 10º C a 35º C (+50 a + 95º F)
Humedad relativa 30% a 75%
Presión ambiente 700 hPa a 1060 hPa
Dimensiones de la base de carga Longitud = 205 x Ancho = 150 x
Altura = 85 (mm)
Peso de la base de carga 250 g
Tiempo de carga apróx. 2 horas (con la batería vacía)
Suministro eléctrico de la pieza de mano batería Li-Po
Máximo tiempo de batería apróx. 60 min. (con una batería
ueva completamente cargada).
Fuente de luz Polywave
®
LED
Franja de longitud de onda 385–515 nm
Intensidad lumínica máx. 1.200 mW/cm
2
± 10%
Funcionamiento 5 min. en funcionamiento /
6 minutos desconectado
(intermitentemente)
Dimensiones de la pieza de mano Longitud = 260 x Ancho = 42 x
Alto = 120 (mm)
Peso de la pieza de mano (batería incl.) 225 g
Condiciones de transporte y almacenamiento:
Temperatura -20º C a +60º C (-4 a 158º F)
Humedad relativa 10% a 75%
Presión ambiente 500 hPa a 1060 hPa
Bluephase N se debe almacenar un estancias cerradas y techadas.
Proteja la lámpara de golpes fuertes.
Batería
No almacenar a temperaturas por encima de 40º C / 104º F
(o 60º C /140º F durante cortos períodos de tiempo). Temperatura
de almacenamiento recomendada 15 – 30 °C (59 – 86 °F).
Almacene siempre la batería cargada
y nunca más de 6 meses
44
Caro cliente,
um endurecimento perfeito constitui a base para
uma restauração de elevada qualidade e de longa
duração de todos os materiais fotopolimerizáveis.
Para esse facto contribui, decisivamente, o
aparelho de polimerização escolhido. Por isso,
constatamos com satisfação que escolheu o
Bluephase N. Trata-se de um produto médico de
elevada qualidade que é fabricado com base nas
normas vigentes e segundo os conhecimentos
técnicos e científicos actuais.
As instruções de Uso explicam como se coloca o
aparelho em funcionamento com segurança, como
se podem aproveitar facilmente todas as suas
possibilidades e como se faz uma boa conservação
para poder usá-lo durante muito tempo.
Estamos à sua disposição para qualquer questão
que nos queira colocar (endereços na contracapa)
A sua equipa Ivoclar Vivadent
45
Vista geral de produtos 46
Lista de peças
Indicações da estação de carregamento
Indicações da peça de mão
Segurança 48
Utilização prevista
– Indicação
– Símbolos
Recomendações relativas à segurança
Colocação em funcionamento 50
Estação de carregamento
Peça de mão
– Bateria
Modo de cabo Click & Cure
Operação 53
Selecção do programa e do tempo de exposição
Função de memória Cure Memory
Iniciar
Sinais acústicos
Intensidade da luz
Medição da intensidade da luz
Manutenção e limpeza 55
– Corpo
Condutor de luz
Contactos da bateria
Eliminação
E se …? 56
Garantia / Procedimento em caso de necessidade de reparação 57
Especificação do produto 57
Índice
Português
46
Lista de peças
1 Peça de mão
2 Botão de ligar
3 Visor
4 Tecla de selecção do programa
5 Tecla de selecção do tempo
6 Tecla de intensidade sonora
7 Indicador de alimentação eléctrica
8 Estação de carregamento
9 Cabo de ligação à rede
10 Unidade de alimentação
11 Bateria
12 Condutor de luz 10 mm
13 Cone anti-encandeamento
10
11
311213
2
9
7 8
5 6 4
47
Indicações da estação de carregamento
Indicações da peça de mão
IIndicador mostra uma luz azul - A estação de
carregamento está ligada à alimentação eléctrica
Programa de exposição ou estado de serviço
Indicação seleccionada ou duração da exposição
Estado de carregamento da bateria
48
Utilização prevista
O Bluephase N é um aparelho de polimerização LED para
a produção de luz azul de alta energia, destinando-se à
polimerização de materiais dentários polimerizáveis
directamente na unidade de tratamento dentária. A
utilização prevista deste aparelho inclui também a
observação das notas e regras que constam destas
instruções de Uso.
Indicação
Com o seu espectro de banda ampla “Polywave
®
”, o
Bluephase N é adequado para a polimerização de todos
os materiais dentários polimerizáveis na faixa de compri-
mentos de onda de 385 a 515 nm. Destes fazem parte
materiais de obturação, bondings/adesivos, forramentos
da base, forramentos, selantes de fissuras, provisórios
bem como materiais de fixação para brackets e peças
técnico-dentárias como, por exemplo, inlays cerâmicos.
Símbolos
Contra-indicação
Símbolos da luz de polimerização
Isolamento duplo (de acordo com Classe II
de segurança)
Proteção contra choque elétrico (aparelho
tipo BF)
Observe as Instruções de Uso
Observe as Instruções de Uso
Advertência
A lâmpada de cura nunca deve ser descarta-
da no lixo doméstico normal. A informação
relativa ao descarte da luz poderá ser
encontrada na homepage local da Ivoclar
Vivadent.
Reciclável
Recomendações relativas à segurança
O Bluephase N é um dispositivo eletrônico e um produto
médico que é sujeito a IEC 60601-1 (EN 60601-1) e às
diretivas da EMC, bem como a Diretiva de Dispositivo
Médico 93/42/EEC. O aparelho está em conformidade
com as normas pertinentes da UE e é classificado como
um produto LED Classe 2.
A lâmpada de cura foi transportada sob condição tecni-
camente segura. Para manter esta condição e assegurar
uma operação segura, todas as notas e os regulamentos
destas Instruções de Uso devem ser observados. Para
poder evitar danos ao equipamento e riscos aos
pacientes, usuários e terceiros, as seguintes instruções de
segurança devem ser obedecidas.
Materiais cuja polimerização é activada fora
da faixa de comprimentos de onda de 385 a
515 nm (actualmente não há conhecimento
deste tipo de materiais). Quando não houver
certeza sobre algum produto, por favor, con-
tacte o fabricante do respectivo material.
Não carregar ou usar o aparelho perto de
substâncias inflamáveis ou combustíveis.
Equipamentos de comunicação de alta
frequência portáteis ou móveis podem inter-
ferir com o equipamento médico. Durante a
operação não é permitido o uso de telefones
móveis.
Segurança
49
Utilização e responsabilidade
O Bluephase N deve ser usado apenas para a finali-
dade para a qual foi projectado. Uma utilização difer-
ente desta ou para além desta é considerada contrária
à utilização prevista. O fabricante não se responsabi-
liza por danos resultantes de qualquer utilização que
não a prevista ou de uma manipulação incorrecta.
Para além disso, o utilizador deverá, ele próprio, veri-
ficar antes da utilização do Bluephase N, se este é uti-
lizável e adequado para os fins previstos. Isto é par-
ti-cularmente importante quando, ao mesmo tempo,
for usado na vizinhança um outro equipamento.
Só podem ser usadas peças sobresselentes e
acessórios originais da Ivoclar Vivadent (ver
acessórios). O fabricante não se responsabiliza por
danos resultantes da utilização de outras peças
sobresselentes ou outros acessórios.
A ponteira de luz é uma parte aplicada e pode
aquecer-se a um máximo de 45°C na interface com a
peça de mão durante a operação.
Tensão de serviço
Antes de ligar o aparelho, verifique se
a) a tensão indicada na placa de características
corresponde à tensão da rede de alimentação e
b) o aparelho adquiriu a temperatura ambiente.
Em caso de manuseamento de bateria ou fonte de
alimentação em separado - por exemplo, durante a
colocação em funcionamento ou o trabalho no modo de
cabo Click & Cure - o contacto com o paciente ou tercei-
ros deve ser evitado. Os contactos eléctricos expostos da
bateria e da ficha de ligação (unidade de alimentação)
não devem ser tocados.
Suspeita de segurança afectada
Caso se suspeite que uma operação segura do aparelho
não é possível, desligue o aparelho da bateria ou da rede
e tome as medidas necessárias para prevenir uma
utilização acidental. Este pode ser o caso, por exemplo,
quando o aparelho está visivelmente danificado ou
quando não está a trabalhar correctamente. O aparelho
só está completamente desligado da rede de alimentação
quando o cabo está desligado da tomada.
Protecção dos olhos
Uma exposição direta ou indireta aos olhos deve ser evitada.
A exposição prolongada à luz torna-se desagradável para
os olhos e pode resultar em dano. Assim, o uso dos
escudos anti-ofuscantes, que são fornecidos, é sempre
recomendado. Os indivíduos que são normalmente
sensíveis à luz, que tomam drogas que são foto-
sensibilizantes, que sofreram cirurgias oculares, ou as
pessoas que trabalham com o aparelho em sua vizinhança,
durante longos períodos de tempo, não devem ficar
expostos à luz deste dispositivo e devem usar óculos de
proteção (laranja), que absorvem luz abaixo de 515 nm.
Bateria
Atenção: Use apenas peças originais, nomeadamente, as
baterias e as estações de carregamento devem ser da
Ivoclar Vivadent.
Não coloque a bateria em curto-circuito. Não guarde a
bateria a temperaturas superiores a 40°C (ou durante um
curto espaço de tempo a 60°C) e guarde-a sempre carre-
gada. Não deve ser guardada mais de 6 meses. Perigo de
explosão quando eliminado no fogo aberto.
Favor observar que a bateria de polímero de lítio pode
reagir com explosão, fogo e desenvolvimento de fumaça,
nos casos de manuseio impróprio e de danos mecânicos.
Baterias de polímero de lítio danificadas não deveriam
mais ser usadas.
Os eletrólitos e vapores de eletrólitos liberados durante a
explosão, o fogo e a fumaça desenvolvida são tóxicos e
corrosivos. No caso de contato com os olhos e pele, lavar
imediatamente com grande quantidade de água. Evitar a
inalação da fumaça. Procurar o médico, em caso de
indisposição.
Formação de calor
Como em todas lâmpadas de potência elevada, a ele-
vada intensidade da luz provoca calor. Uma exposição
prolongada da área próxima da polpa ou do tecido mole
pode provocar danos irreversíveis ou reversíveis. Por isso,
os tempos de exposição devem ser cumpridos. Uma
polimerização contínua superior a 20 segundos na
mesma área e contactos directos com a gengiva, a
mucosa oral e a pele devem ser evitados a todo o custo.
As restaurações indirectas devem ser executadas com
intervalos intermitentes de 20 segundos ou com uma
refrigeração externa por jacto de ar.
Verifique se o aparelho está completo e não sofreu
danos provocados pelo transporte (ver volume de fornec-
imento). Caso faltem peças ou estejam danificadas, con-
tacte imediatamente o vendedor ou o centro de
assistência da sua área.
Estação de carregamento
Antes de ligar o aparelho, verifique se a tensão indicada
na placa de características corresponde à tensão da rede
de alimentação existente. A placa de características
encontra-se no lado inferior da estação de carregamento.
Coloque a estação de carregamento numa superfície
plana adequada.
Remova a folha de protecção da ficha de ligação da uni-
dade de alimentação. Introduza a ficha de ligação da
unidade de alimentação no lado inferior da estação de
carregamento de forma inclinada e exerça uma ligeira
pressão sobre ela até ouvir e sentir que encaixou.
Ligue o cabo de rede à rede de alimentação e à unidade
de alimentação. O indicador de “Power ON” no lado
esquerdo do corpo acende com luz azul (ver indicações
da estação de carregamento).
Peça de mão
Antes da montagem do condutor de luz remova a folha
de protecção da respectiva abertura da peça de mão.
Rode o condutor de luz ligeiramente para introduzi-lo na
peça de mão.
A seguir, coloque o cone anti-encandeamento no
condutor de luz.
50
Colocação em funcionamento
Bateria
Antes da primeira utilização a bateria deve ser total-
mente carregada!
No estado totalmente carregado, a bateria tem uma
capacidade de exposição de aproximadamente 60 minutos.
Introduza a bateria direita na peça de mão até ouvir e
sentir que encaixou.
De um modo suave, colocar a peça de mão no respectivo
descanso da estação de carregamento. Se uma bucha
protetora for usada, é necessário removê-la, antes de
carregar a bateria.
Se possível, use o aparelho sempre com a bateria total-
mente carregada - assim irá prolongar a sua vida útil.
Por isso recomenda-se que coloque a peça de mão na
estação de carregamento após cada paciente. O tempo
de carregamento para baterias vazias é de 2 horas.
Estado de carregamento da bateria
O respectivo estado de carregamento é indicado no visor
com a peça de mão activada, da seguinte forma:
Bateria totalmente carregada
(capacidade de exposição de
aproximadamente 60 minutos)
Bateria meia carregada
Reserva
(no modo de reserva a última barra da
indicação de bateria está vermelha.
A bateria deve ser carregada o mais
rapidamente possível.)
Quando a bateria está sendo carregada a
informação: „Charging Battery“ aparece
brevemente no display, antes do display
retornar ao modo „stand-by“.
Bateria descarregada
Quando a bateria estiver totalmente
descarregada, a peça de mão passa auto-
maticamente para o modo standby. A luz
não acende, e nem o programa nem o
tempo de exposição podem ser ajustados.
Mas a peça de mão pode ser usada em
modo de cabo Click & Cure.
A bateria é uma peça de desgaste que, normalmente,
tem de ser substituída passado aproximadamente
2½ anos. A idade da bateria consta no autocolante na
bateria.
51
52
Modo de cabo Click & Cure
O Bluephase N pode trabalhar sempre e, nomeadamente,
em caso de bateria totalmente descarregada, em modo
de cabo.
Para esse efeito, retire a bateria da peça de mão. A
seguir desligue a unidade de alimentação, tirando a ficha
de li-gação do lado inferior da estação de carregamento.
Não puxe pelo cabo de rede.
Insira a ficha directamente na peça de mão até ouvir e
sentir o encaixe correcto.
Durante a operação a estação de carregamento não pode
carregar a bateria.
O aparelho só estará desligado totalmente da corrente
quando o cabo estiver fora da tomada.
As superfícies contaminadas do aparelho, como os condu-
tores da luz e os cones anti-encandeamento, devem ser
desinfectadas ou autoclavadas, antes de cada utilização.
Além disso, verifique se a intensidade da luz permite uma
polimerização satisfatória. Inspeccione o condutor de luz
regularmente, para detectar eventuais sujidades ou danifi-
cações, e verifique a intensidade da luz (por exemplo, com
o radiómetro “Bluephase Meter” da Ivoclar Vivadent).
Selecção do programa e do tempo
de exposição
O programa e o tempo de exposição podem ser ajustados
individualmente. O Bluephase N dispõe de 3 programas
de exposição que permitem fazer aplicações diferentes.
O programa de exposição pretendido é seleccionado
através dos botões de selecção do programa. O visor
muda o seu aspecto (ver também indicações na peça de
mão).
Da primeira vez que o aparelho é colocado em serviço, ele
irá apresentar a seguinte configuração, pré-ajustada na
fábrica:
HIGH (High Power): 10 segundos
LOW (Low Power): 10 segundos
SOFT (Soft start): 15 segundos
O tempo de exposição pretendido é seleccionado com os
botões de selecção do tempo. Podem ser seleccionados 5,
10, 15, 20 e 30 segundos.
Para seleccionar o tempo de polimerização, consulte o
folheto informativo do material aplicado.
Nos compósitos os tempos de exposição recomendados
referem-se a todas as cores e a camadas com espessuras
máx. de 2 mm, salvo recomendação diferente nos
folhetos informativos do material em questão. Estas
recomendações são aplicáveis, de uma maneira geral, a
exposições em que a janela de emissão da luz do condu-
tor da luz esteja directamente sobre o material que se
pretende polimerizar. Conforme aumenta a distância, os
tempos de exposição têm de ser mais prolongados. Com
uma distância de 9 mm a intensidade efectiva da luz fica
reduzida para aproximadamente 50 %, de modo que o
tempo de exposição recomendado tem de ser duplicado.
Operação
53
Programas de Cura
1) Válido para camadas com espessura máxima de 2 mm contanto que as Instruções de Uso
do respectivo material não registram outra recomendação (poderá ser o caso, p.ex., com as
cores de dentina).
2) Válido para camadas com espessura máxima de 4 mm contanto que as Instruções de Uso
do respectivo material não registram outra recomendação (poderá ser o caso, p.ex., com as
cores de dentina).
3) Válido para camadas com espessura máxima de 3 mm.
4) Válido para fotopolimerização (usar somente a pasta base).
5) Para polimerização dual.
Programa HIGH POWER
Intensidade de luz permanentemente
elevada para a polimerização de
materiais de obturação e fixação
em caso de restaurações directas e
indirectas
Programa SOFT START
Aumento gradual da intensidade de
luz com uma tensão de contracção
reduzida e temperatura mais baixa
para a polimerização de materiais de
obturação
Programa LOW POWER
Intensidade de luz reduzida com
temperatura mais baixa para a polim-
erização de adesivos, liner e materi-
ais de obturação na área
próxima da polpa, para o tratamento
de cavidades da classe V
650 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0 5
650 mW/cm
2
Materiais de restauração
Compósitos
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
Todos os compósitos convencionais
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Compômeros
3)
Compoglass F / Compoglass Flow
Restaurações indiretas /
Materiais de Cimentação
Variolink II Base
4)
/ Variolink Veneer / Variolink N Base
4)
/ Variolink N Clear Veneer / Variolink Ultra
5)
/
Dual Cement
5)
/ Variolink II
5)
Diversos
Helioseal / Helioseal F / Helioseal Clear
Monopaque
MultiCore Flow / Multicore HB
Heliosit Orthodontic
Telio Add-On Flow
Telio Stains
IPS Empress Direct Color
IPS Empress Direct Opaque
Tempo de cura
10 seg.
15 seg.
10 seg.
20 seg.
por mm cerâmica
:
10 seg.
por segmento
10 seg.
20 seg.
20 seg.
10 seg.
15 seg.
10 seg.
10 seg.
20 seg.
Adhesive
AdheSE / AdheSE One F
ExciTE / ExciTE DSC
Tetric N-Bond / Tetric N-Bond Self-Etch
Heliobond (Syntac)
Materiais provisórios
Telio CS Inlay/Onlay
Systemp.inlay/onlay
Fermit / Fermit N
Telio CS Link / Systemp.link
Diversos
Heliosit Orthodontic
Vivaglass Liner
Tempo de cura
10 seg.
10 seg.
10 seg.
10 seg.
20 seg. por segmento
20 seg.
20 seg.
Materiais de restauração
Compósitos
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
Todos os compósitos convencionais
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Compômeros
3)
Compoglass F / Compoglass Flow
Tempo de cura
15 seg.
20 seg.
15 seg.
20 seg.
Intensidade de luz
Tempo de exposição para compósitos
Tempo de exposição para Tetric EvoCeram
/ IPS Empress Direct
1.200 mW/cm
2
15 seg.
10 seg.
54
Função memória e cura
A última regulação usada é guardada automaticamente
na combinação programa e tempo de exposição.
Inicio
A lâmpada é ligada com o botão Iniciar. Decorrido o
tempo de exposição seleccionado, o programa de
exposição termina automaticamente. Se desejar desligar
a luz antes de ter decorrido o tempo fixado, pressione
novamente o botão de ligar. O ventilador é activado
simultaneamente com a luz. No fim do tempo de
exposição, este ainda continua activo, para arrefecer o
aparelho. Enquanto estiver a funcionar o ventilador, a
bateria não deve ser retirada.
Sinais acústicos
Sinais acústicos são emitidos nas seguintes funções:
Iniciar (Parar)
De 10 em 10 segundos
Mudança de programa
Mudança do tempo de exposição
Ligação da bateria à estação de carregamento
Inserção da bateria
Mensagem de erro
Se desejado, o volume de som dos sinais acústicos pode
ser regulado. Para esse efeito, com a luz desligada,
prima o botão azul do nível sonoro (ver Vista geral de
produtos), e com o botão de selecção do tempo (lado
esquerdo) reduza o volume de som ou desligue o som
totalmente. Para ligar os sinais acústicos novamente ou
para aumentar o volume de som, prima o botão azul do
nível sonoro e, a seguir, o botão de selecção do tempo
(lado direito).
Intensidade da luz
A intensidade da luz mantém-se constante durante a
operação do aparelho. No condutor de luz de 10 mm
ornecido, a intensidade da luz foi calibrada em
1200 mW/cm
2
+/- 10%.
O uso do condutor de luz diferente daquele que foi
fornecido tem uma influência direta na intensidade de
luz emitida.
Nos condutores de luz de paredes-paralelas (10 mm), o
diâmetro é igual nas duas extremidades. Nos condutores
de focagem de luz (condutor de luz 10>8 mm, condutor
de luz Pin-Point 6>2 mm), o diâmetro da extremidade
traseira do condutor é maior que aquele da janela de
emissão de luz. Deste modo, a luz azul incidente será
empacotada em uma superfície menor. Isto promove o
aumento da intensidade da luz que será emitida.
Os condutores de luz Pin-Point são adequados para a
cura em áreas confinadas, como na fixação de facetas,
antes da remoção dos excessos. Para completar a poli-
merização, entretanto, será necessário trocar o condutor
de luz.
Medição da intensidade de luz
A intensidade de luz do Bluephase N e do condutor de
luz de 10 mm fornecido pode ser controlada, por exemp-
lo, com o radiómetro “Bluephase Meter”.
Se o valor medido não corresponder à intensidade de luz
esperada:
Verifique o programa de exposição seleccionado
Limpe o sensor óptico do radiómetro que eventual-
mente pode estar sujo
Remova o condutor de luz e limpe a abertura de saída
da luz na peça de mão com um cotonete embebido em
álcool
Limpe o condutor de luz eventualmente sujo (ver
manutenção e limpeza)
Substitua o condutor de luz eventualmente danificado
por um novo
Se estas medidas não produzirem o resultado desejado,
contacte o seu vendedor ou o centro de assistência da
sua área.
55
Por razões de higiene, recomendamos o uso de uma
barreira de proteção descartável para cada paciente.
Certifique-se de ajustar a barreira protetora perfeitamente
à ponteira de luz. Desinfetar superfícies contaminadas do
dispositivo e dos cones anti-dispersão (uma solução
alcoólica, sem aldeídos ou fenóis por ex. FD 366/Dürr
Dental, Incidin Liquid/Ecolab) e esterilize a ponteira de luz
antes de cada uso se luvas protetoras descartáveis não
forem utilizadas. Certifique-se de que líquidos e outras
substâncias estranhas não entrem na peça de mão, no
porta peça de mão e, particularmente, na fonte de energia
durante a limpeza (risco de choque elétrico). Desconecte a
base de recarga da tomada quando for limpála.
Carcaça
Limpe a peça de mão e o suporte da peça de mão com
um lenço com uma solução desinfetante sem aldeído
usual. Não limpe com soluções desinfetantes altamente
agressivas (por exemplo, soluções à base de óleo de
laranja ou com um teor de etanol superior a 40%),
solventes (por exemplo, acetona), ou instrumentos de
pontas, os quais podem danificar ou arranhar o plástico.
Limpe as peças plásticas sujas com uma solução de
sabão.
Ponteira de luz
Antes de limpar e/ou desinfetar a ponteira de luz, ela
deverá ser prétratada. Isso se aplica tanto à limpeza e
desinfecção automática quanto manual.
Pré-tratamento
Remover a contaminação grosseira imediatamente
após a sua utilização, ou no prazo de 2 horas. Para
este fim, enxaguar bem a ponteira de luz em água
corrente (por pelo menos 10 segundos). Como alterna-
tiva, usar uma solução desinfetante sem aldeído ade-
quada para impedir a fixação do sangue.
Para eliminar a contaminação manualmente, use um
pincel ou pano macios. Compósito parcialmente poli-
merizado pode ser removido com álcool e espátula
plástica, se necessário. Não use objetos pontiagudos
ou cortantes, pois podem riscar a superfície.
Limpeza e desinfecção
Para a limpeza, coloque a ponteira de luz em uma solução
de limpeza e certifique-se que ela esteja suficientemente
coberta com líquido (ultra-som ou escove cuidadosamente
com uma escova macia possa potencializar o efeito). Um
agente de limpeza enzimático neutro é recomendado. Ao
limpar e desinfetar, por favor, certifique-se de que os
agentes utilizados não contenham:
ácidos orgânicos, minerais e oxidantes (o valor de
pH mínimo admissível é 5,5)
soluções alcalinas (o valor de pH máximo admissível é
8,5)
agentes oxidantes (por ex. peróxido de hidrogênio)
Depois, remova a ponteira de luz da solução e lave bem
com água corrente fresca (por pelo menos 10 segundos).
Limpar em um desinfetador térmico é uma alternativa
eficaz.
Esterilização
Uma limpeza e desinfecção minuciosa são imperativas
para garantir que a esterilização subsequente seja eficaz.
Utilize apenas esterilização em autoclave para esta finali-
dade. O tempo de esterilização (tempo de exposição à
temperatura de esterilização) é de 4 minutos a 134 ° C, a
pressão deverá ser de 2 bar. Seque a ponteira de luz
esterilizada utilizando o programa de secagem especial
da sua autoclave a vapor ou ar quente. A ponteira de luz
foi testada até 200 ciclos de esterilização.
Depois disso, verifique a ponteira de luz quanto a danos.
Segure-a contra a luz. Se os segmentos individuais apa-
recem em preto, fibras de vidro estão partidas. Se este
for o caso, substitua a ponteira de luz por uma nova.
Contatos da bateria
Para assegurar uma condutividade confiável em todo
momento, mantenha os contatos da bateria sem
qualquer contaminação possível (por ex. resíduos de
compósitos). Para esta finalidade, limpe os contatos
afetados regularmente durante a desinfecção usual com
lenços umedecidos (após cada paciente).
Descarte
A lâmpada de cura não deve ser descartada
como lixo urbano. Descarte as baterias e lâm-
padas de polimerização inutilizáveis de acordo
com os requisitos legais correspondentes em
seu país. Baterias não podem ser incineradas.
Manutenção e limpeza
56
E se …?
Símbolo Causa do problema Solução
Defeito electrónico na peça de
mão ou na bateria
O aparelho está demasiado
quente ou demasiado frio
Defeito electrónico bateria
Bateria descarregada
Os contactos de carregamento
estão sujos.
Bateria em modo de segurança
(circuito protetor integrado).
A unidade de alimentação não
está ligada ou tem algum
defeito.
Remova a bateria e introduza-a novamente. Se o defeito per-
sistir, substitua a bateria pela fonte de alimentação (Click & Cure).
Se o defeito ainda continuar, contacte o seu vendedor ou o
centro de assistência da sua área.
Deixe arrefecer o aparelho (ou se este estiver demasiado frio,
deixe-o adquirir temperatura ambiente), espere algum tempo e
tente novamente. Se o defeito persistir, contacte o seu vendedor
ou o centro de assistência da sua área.
Remova a bateria e introduza-a novamente. Se o aparelho
continuar a indicar um defeito, coloque-o na estação de
carregamento.
Se, mesmo assim, o defeito persistir, substitua a bateria pela
fonte de
alimentação (Click & Cure). Por favor, contacte o seu vendedor ou o
centro de assistência da sua área.
Posicionar a peça de mão na base de carga. A informação:
„Charging Battery“ aparece brevemente no display, antes do
display retornar ao modo „stand-by“.
Se a bateria não carregar, limpe os contactos. Se o defeito
persistir, contacte o seu vendedor ou o centro de assistência da
sua área.
Verifique se a peça de mão se encontra correctamente inserida
na estação de carregamento. Limpe os contactos da bateria.
Informação: „Charging Battery“ aparece brevemente no display,
antes do display retornar ao modo „stand-by“.
Se a limpeza dos contactos da bateria não der os resultados
desejados, retire a bateria do aparelho. Separada da peça de
mão, carregar a bateria na estação de carregamento, durante
aprox. 10 minutos. Se o defeito persistir, contacte o seu vendedor
ou o centro de assistência da sua área.
Verificar se a unidade de alimentação está corretamente posicio-
nada na estação de carregamento (também confira os contatos
de carregamento) ou se a unidade de alimentação está conecta-
da ao suprimento de força através do cabo de força (display da
unidade de alimentação acende a luz verde quando está
trabalhando de modo correto).
Se o defeito persistir, contacte o seu vendedor ou o centro de
assistência da sua área.
O carrega-
mento não
é indicado
no visor.
O LED da
estação de
carrega-
mento não
acende.
57
Garantia / Procedimento em caso
de necessidade de reparação
Especificação do produto
A garantia para o Bluephase N é de 3 anos (bateria 1
ano), a partir da data de compra.
Em caso de avarias provocadas por defeitos de material
ou fabrico, a garantia abrange a reparação gratuita do
aparelho. Para além disso, a garantia não concede o
direito à indemnização de eventuais danos materiais ou
morais. O aparelho deve ser usado apenas para a finali-
dade para a qual foi projectado. Uma utilização diferente
desta ou para além desta é considerada contrária à uti-
lização prevista. Os danos daí resultantes não são
abrangidos pela garantia ou pela responsabilidade civil.
Nestes danos incluem-se, nomeadamente:
Danos provocados por um manuseamento incorrecto,
o que se aplica, nomeadamente, a baterias não cor-
rectamente guardadas (ver Dados Técnicos: Condições
de transporte e armazenagem)
Danos em peças que durante o funcionamento normal
estão sujeitas ao desgaste (por exemplo, a bateria).
Danos provocados por influências exteriores, por
exemplo, impacto, queda ao chão
Danos provocados por uma montagem ou instalação
errada
Danos provocados pela ligação a uma tensão ou
frequência diferente daquela que é indicada na placa
de características.
Danos provocados por modificações e reparações
incorrectas que não foram realizadas por uma entidade
autorizada.
Num caso de garantia, envie o aparelho completo (peça
de mão, estação de carregamento, bateria, cabo de rede
e unidade de alimentação) dentro da embalagem
original, juntamente com a prova de compra, isento
de custos de envio, ao revendedor ou directamente à
Ivoclar Vivadent.
Qualquer tipo de reparação deve ser executado, exclusi-
vamente, por um centro de assistência Ivoclar Vivadent
qualificado. Se o aparelho tiver um defeito que não pode
resolver, contacte o seu vendedor ou o centro de
assistência da sua área (endereços na contracapa). Uma
descrição clara do defeito ou das circunstâncias que
levaram ao defeito facilita a localização da avaria. Por
favor, junte esta descrição ao seu aparelho.
Forma de apresentação
1 estação de carregamento, 1 cabo de rede, 1 unidade de alimen-
tação, 1 peça de mão, 1 bateria (LiPo), 1 condutor de luz de 10 mm,
3 cones anti-encandeamento, 1 conjunto de instruções de Uso
Acessórios
Para o seu Bluephase N existem os seguintes acessórios:
REF Designação
656196 Condutor de luz 10 mm preto (N)
608538 Condutor de luz Pin-Point 6/2 mm preto
551756 (Cones) anti-encandeamento
592496 Escudo anti-encandeamento
659956 Bateria Bluephase N
659681 Peça de mão Bluephase N
607922 Radiómetro Bluephase Meter
608554 Invólucros de protecção Bluephase
Dados técnicos
Tensão de serviço estação de carregamento 5 VDC
Tensão de serviço peça de mão 3.7 VDC com bateria
5 VDC com fonte de alimentação
Unidade de alimentação 100–240 VAC / 50-60 Hz / máx. 0.4 A
Fabricante Friwo
Saída 5 VDC
Tipo 15.2630
Condições de operação
Temperatura +10 °C a +35 °C
Humidade relativa 30 % a 75 %
Pressão atmosférica 700 hPa a 1060 hPa
Dimensões da estação de carregamento C = 205 mm; L = 150 mm;
A = 85 mm
Peso da estação de carregamento 250 g
Tempo de bateria aprox. 2 horas (com bateria
descarregada)
Alimentação eléctrica peça de mão bateria LiPo
Tempo de exposição máx. aprox. 60 minutos
(com bateria nova e totalmente
carregada)
Fonte de luz „Polywave
®
“ LED
Faixa de comprimentos de onda 385 – 515 nm
Intensidade de luz máx. 1.200 mW/cm
2
± 10 %
Operação 5 min. ligado /
6 min. desligado (intermitente)
Dimensões da peça de mão C = 260 mm; L = 42 mm,
A = 120 mm
Peso da peça de mão (inc. bateria) 225 g
Condições de transporte e armazenagem:
Temperatura -20 °C a +60 °C
Humidade relativa 10 % a 75 %
Pressão atmosférica 500 hPa a 1060 hPa
Guarde o Bluephase N em ambientes fechados ou cobertos.
Não exponha o aparelho a fortes vibrações.
Bateria
não guardar a temperaturas superiores a 40°C (ou durante um
curto espaço de tempo a 60°C), recomenda-se a conservação a
15 – 30 °C.
sempre carregada
e não mais que 6 meses
58
Say›n müflterimiz,
Optimum polimerizasyon, sürekli olarak yüksek
kaliteli restorasyonlar elde etmek için ›fl›kla sert-
leflen tüm materyallerde önemli bir gerekliliktir.
Seçilen polimerizasyon ›fl›¤› da bu anlamda önemli
bir role sahiptir. Bu nedenle Bluephase N cihaz›n›
sat›n ald›¤›n›z için size teflekkür etmek isteriz.
Bluephase N, geçerli endüstriyel standartlarla
uyumlu olarak en son bilim ve teknoloji standart-
lar›na göre tasarlanm›fl yüksek kaliteli bir t›bbi
cihazd›r.
Bu çal›flt›rma talimatlar› cihaz› güvenli flekilde
açman›za, performans›ndan tam olarak
yararlanman›za ve cihaz› uzun süre kullanman›za
yard›mc› olacakt›r.
Baflka sorular›n›z varsa size memnuniyetle yard›mc›
oluruz (adresler için arka sayfaya bak›n).
Ivoclar Vivadent Ekibiniz
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 1
59
Türkçe
Ürüne genel bak›fl 60
Parçalar›n listesi
fiarj ünitesi üzerindeki göstergeler
El tabancas› üzerindeki göstergeler
Güvenlik 62
Amaçlanan kullan›m
Kullan›m alanlar›
‹flaretler ve semboller
Güvenlik notlar›
Cihaz›n aç›lmas› 64
fiarj ünitesi
El tabancas›
Akü
Click & Cure kablolu çal›flt›rma
Çal›flt›rma 67
Polimerizasyon program›n›n ve polimerizasyon süresinin seçilmesi
Cure Memory veri saklama fonksiyonu
Bafllatma
Sesli sinyaller
Ifl›k fliddeti
Ifl›k fliddetinin ölçülmesi
Bak›m ve temizleme 69
Gövde
Ifl›k problar›
Akü uçlar›
Cihaz›n at›lmas›
Bu cihaz aç›lmad›¤›nda …? 70
Garanti 71
Tamir prosedürü 71
Ürün tan›m› 71
‹çindekiler
Türkçe
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 2
60
Parçalar›n listesi
1 El tabancas›
2 Bafllatma dü¤mesi
3 Ekran
4 Program seçim tuflu
5 Zaman seçim tuflu
6 Ses seviyesi tuflu
7 Güç göstergesi
8 fiarj ünitesi
9 Elektrik kablosu
10 Adaptör
11 Akü
12 Ifl›k probu, 10 mm
13 Yans›may› önleyici koni
10
11
31
1213
2
9
7 8
5 6 4
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 3
61
fiarj ünitesi üzerindeki göstergeler
El tabancas› üzerindeki göstergeler
Gösterge mavi renkte yan›yor -
Gösterge ünitesi elektrik beslemesine ba¤l›
Polimerizasyon program› ve çal›flma durumu
Seçilen göstergeler ve polimerizasyon süresi
Akünün flarj durumu
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 4
62
Amaçlanan kullan›m
Bluephase N, yüksek enerjili mavi ›fl›k üreten bir LED poli-
merizasyon ›fl›¤›d›r. Bu cihaz, difl hekimli¤i muayenehane-
sinde do¤rudan kullan›lan ›fl›kla sertleflen dental
materyallerin polimerizasyonunda kullan›l›r. Cihaz›n
amaçlanan kullan›m› ayn› zamanda bu Kullan›m Talimat-
lar›nda yer alan bilgilere uyulmas›n› da kapsar.
Kullan›m alanlar›
“ploywave” genifl bantl› spektrumuyla Bluephase N,
385-515 nm dalga uzunlu¤u aral›¤›nda ›fl›kla sertleflen
tüm dental materyallerin polimerizasyonu için uygundur.
Bu materyaller, restorasyon materyalleri, bonding ajan-
lar›/adezivleri, bazlar›, linerlar›, fissür örtücüler, geçici
dolgular› ve braketler için yap›flt›rma materyalleri ve por-
selen inlayler gibi difl laboratuvar parçalar›n› kapsar.
‹flaretler ve semboller
‹zin verilmeyen kullan›m
Cihazdaki semboller
‹ki kez izole edilmifl
(güvenlik s›n›f› II ile uyumlu cihaz)
Elektrik çarpmas›na karfl› koruma
(BF tipi cihaz)
Çal›flt›rma talimatlar›na uyun
Çal›flt›rma talimatlar›na uyun
Dikkat
Cihaz normal ev çöpüyle birlikte
at›lmamal›d›r. Cihazın giderilmesiyle ilgili
bilgileri kendi ulusal Ivoclar Vivadent ana
sayfanızda bulabilirsiniz.
Geri Dönüfltürülebilir
Güvenlik notlar›
Bluephase N, IEC 60601-1 (EN 60601-1) ve EMC
direktiflerine ve aynı zamanda 93/42/EEC sayılı Tıbbi
Cihaz Direktifi'ne tabi olan bir elektronik cihaz ve tıbbi
üründür. Cihaz, ilgili AB düzenlemeleri ile uyumludur ve
LED Sınıf 2 ürün olarak sınıflandırılmaktadır.
Polimerizasyon ıflık cihazı, güvenli ve teknik açıdan
sa¤lam bir durumda fabrikadan çıkmıfltır. Bu durumu
korumak ve risksiz bir çalıflma sa¤lamak için bu Kullanma
Talimatındaki bilgilere ve kurallara uyulmalıdır.
Ekipmanın zarar görmesini ve hastalar, kullanıcılar ve
üçüncü taraflar açısından riskleri önlemek için afla¤ıdaki
güvenlik talimatları izlenmelidir.
Polimerizasyonlar›n›n 385–515 nm dalga
uzunlu¤u aral›¤›n›n d›fl›nda polimerize olan
materyaller (ancak bugüne kadar bu tip bir
materyal bilinmemektedir). Emin olmad›¤›n›z
ürünlerde, bu ürünlerin imalatç›lar›ndan bilgi
alman›z tavsiye edilir.
Cihazı alev alıcı veya yanıcı maddelerin
yakınında flarj etmeyin ve kullanmayın.
Tafl›nabilir ve mobil yüksek frekansl› iletiflim
sistemleri t›bbi cihazlarda enterferansa neden
olabilir. Bu nedenle cihaz çal›flt›r›ld›¤› s›rada
cep telefonu kullan›lmamal›d›r.
Güvenlik
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 5
63
Kullan›m ve sorumluluk kapsam›
Bluephase N yaln›zca amaçland›¤› flekilde
kullan›lmal›d›r. Baflka bir kullan›m ya da amac›n
d›fl›nda kullan›m, kurallara uygun olarak kabul edilme-
mektedir. Kurallara uygun olmayan kullan›m veya
amaç d›fl› kullan›m sonucunda meydana gelen zarar-
lardan imalatç› sorumlu de¤ildir.
Kullan›c›, Bluephase N’i kullanmadan önce, cihaz›n
öngörülen amaçlar için uygun oldu¤unu ve bu alan-
larda kullan›labildi¤ini kendi sorumlulu¤u dahilinde
kontrol etmekle yükümlüdür. Bu, özellikle cihaz›n
hemen yak›n›nda ayn› anda baflka cihazlar çal›flt›r›ld›¤›
zaman geçerlidir.
Yaln›zca orijinal Ivoclar Vivident yedek parçalar›n› ve
aksesuarlar›n› kullan›n (bkz. Aksesuarlar bölümü). Bafl-
ka yedek parça ve aksesuarlar›n kullan›lmas›ndan
do¤an zararlardan imalatç› sorumlu de¤ildir.
Iflık probu uygulamalı bir parçadır ve kullanım
sırasında el aleti arayüzünde maksimum 45°C
sıcaklı¤a kadar ısınabilir.
Çal›flt›rma gerilimi
Cihaz› açmadan önce afla¤›dakilerden emin olun:
a) Tip plakas›nda belirtilen gerilim besleme flebekesindeki
gerilimle ayn› olmal› ve
b) cihaz ortam s›cakl›¤›na gelmelidir.
Akü veya adaptör (pil paketi) ayr› olarak kullan›ld›¤›nda
(örn. cihaz›n bafllat›lmas› veya Click & Cure iflleviyle
kablolu çal›flt›rma s›ras›nda), cihaz›n hasta veya üçüncü
kiflilerle temas etmesi önlenmelidir. Pilin aç›kta kalan
kontaklar›na veya ba¤lant› fifline (adaptör) dokunmay›n.
Güvenli¤in k›s›tlanmas› ihtimali
Tehlikesiz bir çal›flt›rman›n art›k mümkün olmad›¤›
düflünüldü¤ünde, cihaz›n fifli prizden çekilmeli, akü ile
ba¤lant›s› kesilmeli ve yanl›fll›kla kullan›lmas›na karfl›
emniyete al›nmal›d›r. Gözle görülür bir hasar olmas› veya
do¤ru flekilde çal›flmamas› durumunda, cihaz›n
çal›flt›r›lma güvenli¤i azalabilir. Elektrik flebekesinden
tümüyle ay›rma ancak fiflin prizden çekilmesiyle
mümkündür.
Gözlerin korunmas›
Gözlere dolayl› veya do¤rudan ›fl›n gelmesi önlenmelidir.
Uzun süreli ›fl›n tutulmas› gözler için rahats›zl›k vericidir
ve gözlerde hasarlara neden olabilir. Bu nedenle, ürünle
birlikte verilen yans›ma önleyicinin kullan›lmas› tavsiye
edilir. Ifl›¤a karfl› genel olarak hassas olan, ›fl›¤a karfl›
hassasiyet yüzünden ilaçlar ya da fotosensitizan ilaçlar
alan, göz ameliyat› geçirmifl veya uzun süredir bu cihazla
veya cihaz›n yak›n›nda çal›flan kifliler cihaz›n ›fl›nlar›na
maruz kalmamal› ve 515 nm dalga uzunlu¤u alt›ndaki
›fl›nlar› absorbe eden koruyucu gözlük (turuncu)
takmal›d›r.
Akü
Dikkat: Sadece orijinal parçalar, özellikle I
voclar Vivadent
aküleri ve flarj ünitelerini kullan›n.
Aküye k›sa ba¤lant› yapmay›n. Cihaz› 40°C (veya k›sa bir
süre için 60°C) üzerindeki s›cakl›klarda saklamay›n.
Aküler her zaman flarj edilmifl halde saklanmal›d›r.
Saklama süresi 6 ay› aflmamal›d›r. Aç›k aleve at›lmas›
halinde patlama tehlikesi vard›r.
Lityum polimer pilin uygun olmayan muamele ve mekanik
hasar durumunda patlama, yangın çıkması ve duman
oluflması fleklinde reaksiyon gösterebilece¤ine dikkat
edin. Hasarlı lityum polimer piller artık kullanılmamalıdır.
Patlama, yangın çıkması ve duman oluflması durumunda
serbest kalan elektrolitler ve elektrolit dumanları toksik
ve çürütücüdür. Gözler ve ciltle temas halinde hemen bol
miktarda suyla yıkayın. Dumanları solumaktan kaçının.
Kendinizi iyi hissetmezseniz hemen bir doktora görünün.
S›cakl›k oluflumu
Tüm yüksek performansl› lambalarda oldu¤u gibi, yüksek
›fl›k fliddeti bir s›cakl›¤›n oluflmas›yla ba¤lant›l›d›r. Pulpa
yak›n›ndaki alanlar›n ve yumuflak dokular›n uzun süreli
maruziyetinde geri döndürülemeyen veya döndürülebilen
hasar oluflabilir. Bu nedenle önerilen sertleflme sürelerine
uyulmal›d›r. Ayn› difl yüzeyine 20 saniyeden uzun bir süre
boyunca ›fl›k tutulmas› ve ›fl›¤›n do¤rudan difleti, a¤›z
mukozas› veya deri ile do¤rudan temas etmesinden
kaç›n›lmal›d›r. Gerekirse indirek restorasyonlar›
20 saniyelik aralarla polimerize edin veya hava ak›m›
olan harici so¤utma kullan›n.
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 6
64
Teslimat s›ras›nda içeri¤in eksiksiz olup olmad›¤›n› ve
olas› tafl›ma zararlar› olup olmad›¤›n› kontrol edin (bkz.
teslimat kapsam›). Eksik veya zarar gören parça varsa,
hemen sat›c›n›zla veya Servis Merkezinizle temasa geçin.
fiarj ünitesi
Üniteyi açmadan önce tip plakas› üzerinde belirtilen
gerilimin yerel elektrik beslemesi ile uyumlu oldu¤undan
emin olun. Tip plakas› flarj ünitesinin alt k›sm›nda yer
al›r.
fiarj ünitesini uygun, düz bir masa yüzeyine yerlefltirin.
Adaptörün ba¤lant› fiflindeki koruyucu folyoyu ç›kart›n.
Adaptörün ba¤lant› fiflini flarj ünitesinin alt taraf›ndaki
sokete tak›n. Bunun için fifli hafifçe afla¤›ya e¤in ve yeri-
ne oturdu¤unu duyana ve hissedene kadar hafif bir
bas›nç uygulay›n.
Elektrik kablosunu flebeke prizine ve adaptöre ba¤lay›n.
Cihaz gövdesinin sol taraf›ndaki Güç ON (Aç›k) gösterge-
si mavi renkte yanar (bkz. fiarj ünitesi üzerindeki göster-
geler bölümü).
El tabancas›
Ifl›k probunu monte etmeden önce el tabancas›
üzerindeki ilgili aç›kl›k üzerinde bulunan koruyucu
folyoyu ç›kart›n.
Ifl›k probunu el tabancas›na takt›¤›n›z s›rada hafifçe
döndürün.
Ard›ndan yans›ma önleyiciyi ›fl›k probuna monte edin.
Cihaz›n aç›lmas›
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 7
65
Akü
‹lk kullan›m öncesinde akü tamamen flarj edilmelidir.
Akü tam flarjl› haldeyken yaklafl›k 60 dakikal›k ›fl›nlama
kapasitesine sahiptir.
Aküyü, klik sesiyle yerine oturdu¤unu duyana ve
hissedene kadar do¤rudan el tabancas›n›n içine yerlefltirin.
Kablosuz ıflık cihazınızı flarj bazasındaki yerine
yerlefltiriniz. Bir hijyen koruyucu kullan›l›yorsa, bu
koruyucu akünün flarj edilmesinden önce ç›kar›lmal›d›r.
Mümkün oldu¤u kadar cihaz tamamen dolu akü ile
kullan›lmas› gerekir. Bu, ürünün kullan›m ömrünün
uzamas›na yard›mc› olur. Bu nedenle, her hastadan
sonra el tabancas›n›n flarj ünitesine yerlefltirilmesi
önerilmektedir. Bofl pillerin flarj olma süresi 2 saattir.
Akünün flarj durumu
El tabancas› aç›k oldu¤unda mevcut flarj durumu ekranda
afla¤›daki gibi gösterilir:
Akü tam flarjl›
(yaklafl›k 60 dakikal›k polimeriza
syon
kapasitesi)
Akü yar› flarjl›
Yedek
(Yedek modunda akü göstergesinin son
çubu¤u k›rm›z› renktedir. Akü mümkün
olan en k›sa sürede flarj edilmelidir.)
Pil flarj edilirken, ekran bekleme moduna
dönmeden önce ekranda kısa süre için
„Charging Battery (Pil fiarj Ediliyor)“
görülür.
Akü bofl
Akü tamamen boflald›¤›nda el tabancas›
otomatik olarak bekleme moduna geçer.
Bu durumda ›fl›k çal›flt›r›lamaz ve polime-
rizasyon program› ve süresi ayarlanamaz.
Buna karfl›l›k el tabancas› Click & Cure
kablolu çal›flt›rma moduyla kullan›labilir.
Akü bir sarf malzeme oldu¤undan, yaklafl›k 2½ y›l olan
kullan›m ömrü bittikten sonra de¤ifltirilmelidir. Akünün
ömrü için akü etiketine bak›n.
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 8
66
Click & Cure kablolu çal›flt›rma
Bluephase N’i istedi¤iniz zamanda, özellikle de akünün
tamamen boflald›¤› durumlarda kablolu çal›flt›rma
modunda kullanabilirsiniz. Bunun için aküyü el taban-
cas›ndan ç›kart›n. Ard›ndan flarj ünitesinin alt k›sm›ndaki
ba¤lant› fiflinden adaptörü ç›kart›n. Bu ifllem s›ras›nda
elektrik kablosunu çekmeyin.
Ba¤lant› ucunu ç›t sesi duyacak flekilde el cihaz›na
yerlefltirin.
Kablolu kullan›m s›ras›nda flarj ünitesi, güç kayna¤›na
ba¤l› olmad›¤›ndan ne pili flarj etmede ne de ›fl›k gücü
ölçmede kullan›labilir.
Kablolu çal›flt›rma s›ras›nda flarj ünitesi aküyü flarj
edemez.
Yaln›zca elektrik kablosu soketten çekildi¤i zaman
cihaz›n elektrik ba¤lant›s› kesilir.
Cihaz›n kontamine olmufl yüzeylerini, ›fl›k problar›n› ve
yans›may› önleyici konileri her kulan›m öncesinde
dezenfekte edin veya otokla
vlay›n. Dahas›, ayarlanan ›fl›k
fliddetinin yeterli polimerizasyona olanak sa¤lad›¤›ndan
emin olun. Bunun için düzenli olarak ›fl›k probunu konta-
minasyon ve hasar aç›s›ndan, ayr›ca ›fl›k fliddetini düzenli
aral›klarla kontrol edin (örn. Ivoclar Vivadent radyometresi
"Bluephase Meter” ile).
Polimerizasyon program›n›n ve polimerizasyon
süresinin seçilmesi
Polimerizasyon program› ve süresi ayr› ayr› ayarlanabilir.
Bluephase N, farkl› kullan›m alanlar› için 3 polimerizasy-
on program›na sahiptir. Program seçim tufllar›yla
istedi¤iniz polimerizasyon program›n› seçebilirsiniz. Ekran
buna uygun olarak de¤iflir (ayr›ca bkz. El tabancas›
üzerindeki göstergeler bölümü).
Cihaz ilk kez aç›ld›¤›nda flu ayarlar önceden ayarlanm›fl
olacakt›r:
HIGH (Yüksek güç): 10 saniye
LOW (Düflük güç): 10 saniye
SOFT (Yumuflak bafllang›çl›): 15 saniye
Amaçlanan polimerizasyon süresi zaman seçim tufllar›yla
seçilir. Süre olarak 5, 10, 15, 20 ve 30 saniyeyi
seçebilirsiniz.
Polimerizasyon süresini seçerken kullan›lan materyalin
kullan›m talimatlar›n› dikkate al›n.
Kompozit materyaller için belirtilen polimerizasyon
önerileri tüm renkler için ve, ilgili materyalin kullan›m
talimatlar›nda aksi belirtilmedi¤i sürece, maksimum
2 mm’lik tabaka kal›nl›¤› için geçerlidir. Genellikle bu
öneriler, ›fl›k probunun emisyon penceresinin do¤rudan
polimerize edilecek materyal üzerine yerlefltirildi¤i dur-
umlar için geçerlidir. Ifl›k kayna¤› ile materyal aras›ndaki
mesafenin art›r›lmas›, polimerizasyon süresinin de buna
uygun olarak uzat›lmas›n› gerektirir. Örne¤in, materyale
olan mesafe 9 mm oldu¤unda, etkin ›fl›k fliddeti yaklafl›k
%50 azal›r, böylece önerilen polimerizasyon süresi iki
kat›na ç›kar›lmal›d›r.
Çal›flt›rma
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 9
67
Polimerizasyon program›
1) ‹lgili materyalin Kullanım Talimatı herhangi baflka bir öneride bulunmuyorsa
(örne¤in dentin tonları ile görülebilir) maksimum 2 mm tabaka kalınlı¤ı için
geçerlidir
2) ‹lgili materyalin Kullanım Talimatı herhangi baflka bir öneride bulunmuyorsa
(örne¤in dentin tonları ile görülebilir) maksimum 4 mm tabaka kalınlı¤ı için
geçerlidir
3) Maksimum 3 mm tabaka kalınlı¤ı için geçerlid
ir
4) Iflıkla sertlefltirme için geçerlidir (sadece baz patı kullanımı)
5) Çift sertlefltirme için geçerlidir
HIGH POWER
Direk ve indirek restorasyonlarda
kullan›lan restoratif ve simantasyon
materyallerinin polimerizasyonu için
sürekli yüksek ›fl›k fliddeti.
SOFT START
Restoratif materyallerin
polimerizasyonu için düflük büzülme
stresi ve düflük ›s› geliflimi ile ›fl›k
fliddetinin kademeli olarak
art›r›lmas›
LOW POWER
V. s›n›f kaviteler restore edildi¤i
s›rada pulpa yak›n›ndaki alanlarda
adezivlerin, linerlar›n ve res
toratif
materyallerin polimerizasyonu için
düflük ›s› geliflimi ile düflük ›fl›k
fliddeti
1.200 mW/cm
2
t [s]
05
650 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0
650 mW/cm
2
Restoratif materyaller
Kompozitler
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
Tüm geleneksel kompozitler
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Kompomerler
3)
Compoglass F / Compoglass Flow
‹ndirekt restorasyonlar /
simantasyon materyalleri
Variolink II Base
4)
/ Variolink Veneer / Variolink N Base
4)
/
Variolink N Clear Veneer / Variolink Ultra
5)
/
Dual Cement
5)
/ Variolink II
5)
Çeflitli
Helioseal / Helioseal F / Helioseal Clear
Monopaque
MultiCore Flow / Multicore HB
Heliosit Orthodontic
Telio Add-On Flow
Telio Stains
IPS Empress Direct Color
IPS Empress Direct Opaque
Polimerizasyon süresi
10 s
15 s
10 s
20 s
mm seramik baflına:
10 s, yüzey baflına
10 s
20 s
20 s
10 s
15 s
10 s
10 s
20 s
Adezivler
AdheSE / AdheSE One F
ExciTE / ExciTE DSC
Tetric N-Bond / Tetric N-Bond Self-Etch
Heliobond (Syntac)
Geçici materyaller
Telio CS Inlay/Onlay
Systemp.inlay/onlay
Fermit / Fermit N
Telio CS Link / Systemp.link
Çeflitli
Heliosit Orthodontic
Vivaglass Liner
Polimerizasyon süresi
10 s
10 s
10 s
10 s
20 s yüzey baflına
20 s
20 s
Restoratif materyaller
Kompozitler
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
Tüm geleneksel kompozitler
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
Kompomerler
3)
Compoglass F / Compoglass Flow
Polimerizasyon süresi
15 s
20 s
15 s
20 s
Iflık yo¤unlu¤u
Iflık yo¤unlu¤u kompozitlerin
Iflık yo¤unlu¤u Tetric EvoCeram / IPS Empress Direct
1,200 mW/cm
2
15 s
10 s
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 10
68
Cure Memory veri saklama fonksiyonu
Kullan›lan son ayarlar, polimerizasyon program› ve poli-
merizasyon süresiyle birlikte otomatik olarak kaydedilir.
Bafllatma
Bafllat tufluyla ›fl›¤› aç›n. Seçilen polimerizasyon süresi
doldu¤unda, polimerizasyon program› otomatik olarak
sonland›r›l›r. ‹stendi¤inde, ayarlanan süre dolmadan ›fl›k,
bafllat dü¤mesine bir kez daha bas›larak kapat›labilir.
Ifl›¤›n etkinlefltirilmesiyle ayn› anda fan da aç›l›r. Ifl›klama
süresinin sonland›r›lmas›ndan sonra fan cihaz› so¤utmak
için belirli bir süre çal›flmaya devam eder. Fan çal›flt›¤›
s›rada akü ç›kar›lmamal›d›r.
Sesli sinyaller
Afla¤›daki fonksiyonlarda sesli sinyaller duyulabilir:
Bafllatma (Durdurma)
Her 10 saniyede bir
Program de¤iflikli¤i
Ifl›klama süresi de¤iflikli¤i
Akünün flarj ünitesine ba¤lanmas›
Akünün yerlefltirilmesi
Hata mesaj›
‹stenildi¤inde sesli sinyallerin ses seviyesi ayarlanabilir.
Bunun için ›fl›k kapal› oldu¤u s›rada mavi ses seviyesi
tufluna (bkz. Ürüne genel bak›fl bölümü) bas›n ve sol süre
seçim tuflunu kullanarak ses seviyesini azalt›n veya sesli
sinyalleri tümüyle kesin. Sesli sinyalleri tekrar açmak veya
ses seviyesini art›rmak isterseniz, mavi ses seviyesi tuflu-
na ve ard›ndan da sa¤ süre seçim tufluna bas›n.
Ifl›k fliddeti
Ifl›k fliddeti çal›fla s›ras›nda sabit bir düzeyde tutulur.
Ürünle birlikte verilen 10 mm’lik ›fl›n probu kullan›lacaksa,
›fl›k fliddeti 1200 mW/cm
2
+/- %10 olarak kalibre
edilmifltir.
Ekte verilenin dıflında baflka bir fiber optik kullanılması
durumunda, bunun çıkan ıflı¤ın yo¤unlu¤u üzerinde
do¤rudan etkisi olacaktır.
Paralel duvarlı fiber optiklerde (10 mm) ıflı¤ın girdi¤i pen-
cere ile çıktı¤ı pencerenin çapı aynıdır. Fokus yapan fiber
optiklerde ise (10>8 mm fiber optik, Pin-Point [i¤ne ucu]
fiber optik 6>2 mm) ıflı¤ın girdi¤i pencerenin çapı çıktı¤ı
pencerenin çapından daha büyüktür. Böylece, giren mavi
ıflık daha küçük bir yüzeye odaklanır. Bunun sonucunda
çıkan ıflının yo¤unlu¤u artar.
Pin-Point (i¤ne uçlu) fiber optikler noktasal polimeriza-
syonlarda, örne¤in fazlalıklar alınmadan önce venerin
tesbit edilmesinde, kullanılır. Komple sertlefltirme ifllemi
için fiber opti¤in de¤ifltirilmesi gerekmektedir.
Ifl›k fliddetinin ölçülmesi
Bluephase N ile ürünle birlikte verilen 10 mm’lik ›fl›k
probunun ›fl›k fliddeti Bluephase Meter kullan›larak
kontrol edilebilir.
Ölçülen de¤er beklenen ›fl›k fliddetiyle uyumlu de¤ilse, flu
ifllemleri gerçeklefltirin:
– Seçilen polimerizasyon program›n› kontrol edin.
Radyometrenin muhtemelen kirlenmifl ›fl›k sensörünü
temizleyin.
Ifl›k probunu ç›kart›n ve el tabancas›ndaki ›fl›k emisyon
penceresini alkol ile ›slat›lm›fl bir pamuklu çubukla
temizleyin.
Muhtemelen kirlenmifl ›fl›k probunu temizleyin (bkz.
Bak›m ve temizleme bölümü).
Hasar görmüfl ›fl›k probunu yenisiyle de¤ifltirin.
Al›nan bu önlemler sonuçlar› iyilefltirmezse, lütfen
sat›c›n›zla veya yerel Servis Merkezinizle temasa geçin.
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 11
69
Hijyenik nedenlerden dolayı her hastada tek kullanımlık
bir koruyucu kılıf kullanmanızı tavsiye ederiz. Koruyucu
kılıfın ıflık probuna tam olarak oturdu¤undan emin olun.
Her kullanımdan önce cihazın kontamine olmufl
zeylerini ve yansıma önleyici konileri dezenfekte edin
(FD 366/Dürr Dental, Incidin Liquid/Ecolab) ve tek kul-
lanımlık koruyucu kılıf kullanılmıyorsa, ıflık probunu
sterilize edin. Temizlik sırasında el aletine, flarj ünitesine
ve özellikle flarj aletine herhangi bir sıvı veya baflka
yabancı bir madde girmedi¤inden emin olun (elektrik
çarpması riski). Temizlerken flarj ünitesini güç
kayna¤ından ayırın.
Gövde
El aletini ve el aleti tutucusunu aldehit içermeyen,
alıflılagelmifl bir dezenfektan solüsyonla silin. Plasti¤e
zarar verebilecek veya çizebilecek kuvvetli dezenfektan
solüsyonlarla (örn. portakal ya¤ı esaslı veya etanol içeri¤i
%40'ın üzerinde olan solüsyonlar), çözücülerle (örn.
aseton) veya sivri aletlerle temizlemeyin. Kirli plastik
kısımları sabunlu suyla temizleyin.
Iflık probu
Iflık probunu temizlemeden ve/veya dezenfekte etmeden
önce ön ifllem uygulayın. Bu durum hem otomatik hem
de manuel temizlik ve dezenfeksiyon için geçerlidir.
Ön ifllem
Kaba kirleri kullanımdan hemen sonra veya en geç
2 saat içerisinde giderin. Bu amaçla ıflık probunu akan
su altında iyice durulayın (en az 10 saniye). Alternatif
olarak, kanın kurumasını önlemek için aldehit
içermeyen uygun bir dezenfektan solüsyon kullanın.
Kontaminasyonu manuel olarak gidermek için
yumuflak bir fırça veya yumuflak bir bez kullanın.
Kısmen poli merize olmufl kompozit, alkolle ve gerekirse
bir plastik spatülle çıkarılabilir. Keskin veya sivri
nesneler kullanmayın çünkü bunlar yüzeyi çizebilir.
Temizlik ve dezenfeksiyon
Iflık probunu temizlemek için bir temizleme solüsyonuna
batırın ve yeterince sıvıyla kaplan-masını sa¤layın (ultra-
son veya yumuflak bir fırçayla dikkatlice fırçalamak
solüsyonun etkisini destekleyebilir). Nötr bir enzimatik
temizleme ajanı tavsiye edilir. Temizlerken ve dezenfekte
ederken, lütfen kullanılan ajanların afla¤ıdakileri içerme-
di¤inden emin olun:
Organik, mineral ve oksitleyici asitler (izin verilebilir
minimum pH de¤eri 5,5)
Alkali solüsyon (izin verilebilir maksimum pH de¤eri 8,5)
Oksitleyici ajan (örn. hidrojen peroksit)
Ardından ıflık probunu solüsyondan çıkarın ve akan su
altında iyice durulayın (en az 10 sa-niye). Bir termal
dezenfektörde temizlemek de etkili bir alternatiftir.
Sterilizasyon
Bir sonraki aflamada uygulanacak sterilizasyon iflleminin
etkili olmasını sa¤lamak için iyi bir temizlik ve
dezenfeksiyon flarttır. Bu amaçla sadece otoklav
sterilizasyonu uygulayın. Ste-rilizasyon süresi
(sterilizasyon sıcaklı¤ındaki maruziyet süresi) 134°C
sıcaklıkta 4 dakikadır; basınç 2 bar olmalıdır. Sterilize
edilmifl ıflık probunu buhar otoklavınızın özel kurutma
programını veya sıcak hava kullanarak kurutun. Iflık
probu 200 sterilizasyon döngüsüne kadar test edilmifltir.
Bundan sonra, ıflık probunu hasar açısından kontrol edin.
Iflı¤a do¤ru tutun. E¤er kimi bölümler siyah görünüyorsa,
cam fiberler kırılmıfltır. Bu durumda, ıflık probunu yenisiy-
le de¤ifltirin.
Pil kontakları
Her zaman güvenilir iletkenlik sa¤lamak için pil
kontaklarını olası kontaminasyonlardan koruyun (örn.
kompozit kalıntısı). Bu amaçla, rutin olarak uygulanan
silerek dezenfeksiyon ifllemi sırasında (her hastadan
sonra) etkilenen kontakları düzenli olarak temizleyin.
Bertaraf etme
Polimerizasyon lambası evsel atık olarak
bertaraf edilmemelidir. Kullanılamayacak
durumdaki pilleri ve polimerizasyon
lambalarını ülkenizdeki ilgili yasal
gerekliliklere göre bertaraf edin. Piller
yakılmamalıdır.
Bak›m ve temizleme
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 12
70
Bu cihaz aç›lmad›¤›nda ...?
Sembol Sorunun nedeni Sorun giderme
El tabancas›nda veya aküde
elektronik ar›za
Cihaz afl›r› ›s›nd› veya afl›r›
so¤udu
Aküde elektronik ar›za
Akü bofl
fiarj uçlar› kontamine oldu
Akü güvenlik modunda
(entegre koruyucu devre)
Adaptör ba¤l› de¤il veya ar›zal›
Aküyü ç›kar›p yeniden yerlefltirin. Hata devam ederse, akünün
yerine adaptörü (Click & Cure) kullan›n.
Buna ra¤men hata devam ederse, lütfen sat›c›n›zla veya yerel
Servis Merkezinizle temasa geçin.
Cihaz›n so¤umas›n› bekleyin (veya cihaz afl›r› so¤umuflsa oda
s›cakl›¤›na gelmesini bekleyin) ve belirli bir süre sonra tekrar
deneyin. Hata devam ederse, lütfen sat›c›n›zla veya yerel Servis
Merkezinizle temasa geçin.
Aküyü ç›kar›p yeniden yerlefltirin. Ar›za göstergesi silinmiyorsa,
cihaz› flarj ünitesine tak›n.
Buna ra¤men hata devam ederse, akünün yerine adaptörü
(Click & Cure) kullan›n.
El tabancasını flarj tabanında konumlandırın. Ekran bekleme
moduna dönmeden önce ekranda kısa süre için „Charging Battery
(Pil fiarj Ediliyor)“ görülür. Akü flarj olmuyorsa, uçlar temizlenmeli-
dir. Hata devam ederse, lütfen sat›c›n›zla veya yerel Servis
Merkezinizle temasa geçin.
El tabancas›n›n flarj ünitesine do¤ru yerlefltirilip yerlefltirilmedi¤imi
kontrol edin. Akü uçlar›n› temizleyin.
Bilgi: Ekran bekleme moduna dönmeden önce ekranda kısa süre
için „Charging Battery (Pil fiarj Ediliyor)“ görülür.
Akü uçlar›n›n temizlenmesine ra¤men hata devam ederse, aküyü
cihazdan ç›kart›n. Pili ıflık cihazınızdan ayrı olarak flarj bazasında
yaklaflık 10 dakika flarj ediniz.
Hata devam ederse, lütfen sat›c›n›zla veya yerel Servis
Merkezinizle temasa geçin.
Güç ünitesinin flarj bazasına do¤ru olarak konumlandırıldı¤ını
(flarj temas alanlarıda kontrol edilmelidir) veya güç ünitesinin güç
kayna¤ına kablo ile ba¤lı olup olmadı¤ını (do¤ru ba¤lanmıflsa
güç ünitesindeki göstergede yeflil ıflık yanacaktır) kontrol edin.
Hata devam ederse, lütfen sat›c›n›zla veya yerel Servis Merkezi-
nizle temasa geçin.
fiarj
s›ras›nda
ekranda
gösterge
yok.
fiarj
ünitesindeki
LED
yanm›yor.
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 13
71
Garanti /
Tamir prosedürü
Ürün tan›m›
Bluephase N garantisi sat›n alma tarihinden itibaren
3 y›ld›r (akü için 1 y›l).
Materyal veya üretim hatas›ndan kaynaklanan ar›zalarda
cihaz garanti süresi dahilinde ücretsiz tamir edilir.
Garanti, belirtilen hasarlar d›fl›nda maddi veya maddi
olmayan herhangi bir hasar›n tazmin edilmesi hakk›n›
sa¤lamaz. Cihaz yaln›zca amaçland›¤› flekilde
kullan›lmal›d›r. Farkl› veya amac›n› aflan kullan›ma izin
verilmez. Yanl›fl kullan›mdan kaynaklanan zararlar ve
bunlar›n garanti kapsam›nda tazmini için üretici herhangi
bir sorumluluk kabul etmez. Bu, özellikle afla¤›dakiler için
geçerlidir:
Amaca uygun olmayan kullan›mdan, özellikle do¤ru
flekilde saklanmayan akülerden kaynaklanan zarar
(bkz. Teknik veriler: Tafl›ma ve saklama koflullar›).
Bileflenlerde, normal çal›flt›rma s›ras›ndaki afl›nmadan
kaynaklanan hasar (örn. akü).
Darbe veya yere düflürülmesi gibi d›fl etkilerden
kaynaklanan hasar.
Hatal› ayar veya kurulumdan kaynaklanan hasar.
Cihaz›n, tip plakas›nda belirtilenden farkl› voltaj ve
frekansa sahip bir flebekeye ba¤lanmas›ndan
kaynaklanan hasar.
Yetkili olmayan servis merkezleri taraf›ndan yap›lan
uygun olmayan tamirler ve modifikasyonlardan
kaynaklanan hasar.
Garanti durumunda cihaz›n tamam› (el tabancas›, flarj
ünitesi, akü, elektrik kablosu ve adaptör) sat›fl fifliyle
birlikte, orijinal ambalaj›nda ve ilgili karton malzemelerle,
navlun ödenmifl olarak teslimat deposuna veya do¤rudan
Ivoclar Vivadent’e gönderilmelidir.
Her türlü tamir ifli yaln›zca yetkili Ivoclar Vivadent Servis
Merkezleri taraf›ndan gerçeklefltirilmelidir.
Gideremedi¤iniz bir ar›za durumunda lütfen sat›c›n›zla
veya Servis Merkezinizle temasa geçin (adresler arka
sayfadad›r). Ar›zan›n veya ar›zan›n ortaya ç›kt›¤›
koflullar›n net bir dille aç›klamas› sorunun tespitini
kolaylaflt›racakt›r. Lütfen cihaz› iade ederken bu
aç›klamay› da ekleyin.
Teslimat kapsam›
1 fiarj ünitesi, 1 Elektrik kablosu, 1 Adaptör, 1 El tabancas›,
1 Akü (Li-Po), 1 10 mm’lik ›fl›k probu, 3 Yans›may› önleyici koni, 1 Kullan›m
Talimat›
Aksesuar
Bluephase N cihaz›n›z için afla¤›daki aksesuarlar mevcuttur:
REF Aç›klama
656196 Ifl›k probu, 10 mm siyah (N)
608538 Nokta hedefli ›fl›k probu, 6/2 mm siyah
551756 Yans›ma önleyici (koni)
592496 Yans›may› önleyici plaka
659956 Bluephase N akü
659681 Bluephase N el tabancas›
607922 Bluephase Meter
608554 Bluephase koruyucu kolluklar
Teknik veriler
‹flletme gerilimi flarj ünitesi 5 VDC
‹flletme gerilimi e
l tabancas› 3.7 VDC, akü ile
5 VDC, adaptör ile
Adaptör 100–240 VAC / 50-60 Hz / maks. 0.4 A
Üretici: Friwo
Ç›k›fl 5 VDC
T‹P 15.2630
Çal›flma koflullar›
S›cakl›k +10°C ila +35°C
Ba¤›l nem %30 ila %75
Hava bas›nc› 700 hPa ila 1060 hPa
Boyutlar flarj ünitesi U = 205 mm; G = 150 mm; Y = 85 mm
A¤›rl›k flarj ünitesi 250 g
fiarj süresi yakl. 2 saat (bofl aküde)
Elektrik beslemesi el tabancas› Li-Po akü
Maks. ›fl›klama süresi
yakl. 60 dak. (yeni, tam flarjl› akü ile)
Ifl›k kayna¤› Polywave
®
LED
Dalga boyu aral›¤› 385–515 nm
Ifl›k fliddeti maks. 1.200 mW/cm
2
± %10
‹flletme 5 dak. aç›k /
6 dak. kapal› (aral›kl› olarak)
Boyutlar El tabancas›: U = 260 mm; G = 42 mm; Y = 120 mm
A¤›rl›k el tabancas› (akü dahil) 225 g
Tafl›ma ve saklama koflullar›:
S›cakl›k -20°C ila +60°C
Ba¤›l nem %10 ila %75
Hava bas›nc› 500 hPa ila 1060 hPa
Bluephase N kapa
l› veya üzeri örtülü mekanlarda saklanmal›d›r.
Cihaz› güçlü sars›nt›lara maruz b›rakmay›n.
Akü
40 °C (veya k›sa bir süreli¤ine 60 °C) üzerindeki s›cakl›klarda
saklamay›n. Önerilen saklama s›cakl›¤›: 15 – 30 °C.
Aküyü her zaman flarjl› halde ve en fazla 6 ay boyunca saklay›n.
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 14
72
삇ʇÂÏ˚È ÔÓÍÛÔ‡ÚÂθ,
éÔÚËχθ̇fl ÔÓÎËÏÂËÁ‡ˆËfl fl‚ÎflÂÚÒfl
ÓÒÌÓ‚ÓÈ ıÓÓ¯Â„Ó Í‡˜ÂÒÚ‚‡ ÂÒÚ‡‚‡ˆËË
ËÁ β·˚ı Ò‚ÂÚÓÓÚ‚Âʉ‡ÂÏ˚ı χÚÂˇ-
ÎÓ‚. ëÛ˘ÂÒÚ‚ÂÌÌÛ˛ Óθ ÔË ˝ÚÓÏ Ë„‡ÂÚ
Ô‡‚ËθÌ˚È ‚˚·Ó ÔÓÎËÏÂËÁ‡ˆËÓÌÌÓÈ
·ÏÔ˚. èÓ˝ÚÓÏÛ Ï˚ Ó˜Â̸ ‡‰˚, ˜ÚÓ Ç˚
ÓÒÚ‡ÌÓ‚ËÎË Ò‚ÓÈ ‚˚·Ó ̇ ·ÏÔ Bluephase
N. ùÚ‡ ·ÏÔ‡ Ô‰ÒÚ‡‚ÎflÂÚ ÒÓ·ÓÈ ‚˚ÒÓÍÓ-
͇˜ÂÒÚ‚ÂÌÌ˚È ÔÓ‰ÛÍÚ Ï‰ˈËÌÒÍÓÈ ÚÂıÌË-
ÍË, ÍÓÚÓ˚È ·˚Î ÒÍÓÌÒÚÛËÓ‚‡Ì ‚
ÒÓÓÚ‚ÂÚÒÚ‚ËË Ò ‰ÂÈÒÚ‚Û˛˘ËÏË ÌÓχÏË,
‡ Ú‡ÍÊÂ Ò Û˜ÂÚÓÏ ÒÓ‚ÂÏÂÌÌÓ„Ó ÛÓ‚Ìfl
Á̇ÌËÈ Ë ÚÂıÌËÍË.
Ç ËÌÒÚÛ͈ËË Í ÔË·ÓÛ Ç˚ ̇ȉÂÚ ÔÓ‰-
Ó·ÌÓ ‡Á˙flÒÌÂÌËÂ, Í‡Í Â„Ó ÔËÏÂÌflÚ¸, ͇Í
̇˷ÓΠÔÓÒÚ˚Ï Ë Û‰Ó·Ì˚Ï ÒÔÓÒÓ·ÓÏ
ËÒÔÓθÁÓ‚‡Ú¸ ‚ÒÂ Â„Ó ‚ÓÁÏÓÊÌÓÒÚË Ë Í‡Í Á‡
ÌËÏ Ûı‡ÊË‚‡Ú¸, ˜ÚÓ·˚ ÓÌ ÔÓÒÎÛÊËΠLJÏ
Í‡Í ÏÓÊÌÓ ‰Óθ¯Â.
ÖÒÎË Û Ç‡Ò ‚ÓÁÌËÍÌÛÚ ‚ÓÔÓÒ˚, Ï˚ Ò
Û‰Ó‚ÓθÒÚ‚ËÂÏ Ì‡ ÌËı ÓÚ‚ÂÚËÏ (‡‰ÂÒ‡
Ç˚ ̇ȉÂÚ ̇ ÔÓÒΉÌÂÈ ÒÚ‡ÌˈÂ
ËÌÒÚÛ͈ËË).
àÒÍÂÌÌ LJ¯
Ivoclar Vivadent
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 15
73
English
êyccÍËÈ
é·ÁÓ ÔÓ‰ÛÍÚ‡ 74
– èÂ˜Â̸ ÒÓÒÚ‡‚Ì˚ı ˜‡ÒÚÂÈ
– èÓ͇Á‡ÌËfl Á‡fl‰ÌÓ„Ó ·ÎÓ͇
– èÓ͇Á‡ÌËfl ̇ÍÓ̘ÌË͇
ÅÂÁÓÔ‡ÒÌÓÒÚ¸ 76
– àÒÔÓθÁÓ‚‡ÌË ÔÓ Ì‡Á̇˜ÂÌ˲
– èÓ͇Á‡ÌËfl
– ê‡Á˙flÒÌÂÌË Á̇ÍÓ‚
– íÂıÌË͇ ·ÂÁÓÔ‡ÒÌÓÒÚË
Ç‚Ó‰ ‚ ˝ÍÒÔÎÛ‡Ú‡ˆË˛ 78
– á‡fl‰Ì˚È ·ÎÓÍ
– ç‡ÍÓ̘ÌËÍ
– ÄÍÍÛÏÛÎflÚÓ
– ꇷÓÚ‡ ˜ÂÂÁ ͇·Âθ Click & Cure
àÒÔÓθÁÓ‚‡ÌË ÔË·Ó‡ 81
– Ç˚·Ó ÔÓ„‡ÏÏ˚ ÔÓÎËÏÂËÁ‡ˆËË Ë ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË
– îÛÌ͈Ëfl ÒÓı‡ÌÂÌËfl ‰‡ÌÌ˚ı Cure Memory
– á‡ÔÛÒÍ
– á‚ÛÍÓ‚˚ Ò˄̇Î˚
– åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl
– àÁÏÂÂÌË ÏÓ˘ÌÓÒÚË Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl
íÂıÓ·ÒÎÛÊË‚‡ÌËÂ Ë Ó˜ËÒÚ͇ 83
– äÓÔÛÒ
– ë‚ÂÚÓ‚Ó‰
– äÓÌÚ‡ÍÚ˚ ‡ÍÍÛÏÛÎflÚÓ‡
– ìÚËÎËÁ‡ˆËfl
óÚÓ ‰Â·ڸ, ÂÒÎË …? 84
ɇ‡ÌÚËfl / ÑÂÈÒÚ‚Ëfl ‚ ÒÎÛ˜‡Â ÂÏÓÌÚ‡ 85
ëÔˆËÙË͇ˆËfl 85
ëÓ‰ÂʇÌËÂ
êÛÒÒÍËÈ
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 16
74
èÂ˜Â̸ ÒÓÒÚ‡‚Ì˚ı ˜‡ÒÚÂÈ
1 ç‡ÍÓ̘ÌËÍ
2 äÌÓÔ͇ ÒÚ‡Ú‡
3 ÑËÒÔÎÂÈ
4 ä·‚˯‡ ‚˚·Ó‡ ÔÓ„‡ÏÏ˚
5 ä·‚˯‡ ‚˚·Ó‡ ‚ÂÏÂÌË
6 ä·‚˯‡ „ÛÎËÓ‚ÍË „ÓÏÍÓÒÚË
7 é·ÓÁ̇˜ÂÌË ÏÓ˘ÌÓÒÚË
8 á‡fl‰Ì˚È ·ÎÓÍ
9 ëÂÚ‚ÓÈ Í‡·Âθ
10 ëÂÚ‚ÓÈ ·ÎÓÍ
11 ÄÍÍÛÏÛÎflÚÓ
12 ë‚ÂÚÓ‚Ó‰ 10 ÏÏ
13 ᇢËÚÌ˚È ˝Í‡Ì
10
11
31
1213
2
9
7 8
5 6 4
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 17
75
èÓ͇Á‡ÌËfl Á‡fl‰ÌÓ„Ó ·ÎÓ͇
èÓ͇Á‡ÌËfl ̇ÍÓ̘ÌË͇
ÉÓËÚ ÒËÌËÈ Ó„ÓÌÂÍ -
á‡fl‰Ì˚È ·ÎÓÍ ÔÓ‰Íβ˜ÂÌ Í ˝ÎÂÍÚÓÒÂÚË.
- èÓ„‡Ïχ ÔÓÎËÏÂËÁ‡ˆËË ËÎË ÂÊËÏ
‡·ÓÚ˚ ÔË·Ó‡
- Ç˚·‡ÌÌ˚ ÔÓ͇Á‡ÌËfl ËÎË ‰ÎËÚÂθÌÓÒÚ¸
ÔÓÎËÏÂËÁ‡ˆËË
- ìÓ‚Â̸ Á‡fl‰ÍË ‡ÍÍÛÏÛÎflÚÓ‡
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 18
76
ÅÂÁÓÔ‡ÒÌÓÒÚ¸
àÒÔÓθÁÓ‚‡ÌË ÔÓ Ì‡Á̇˜ÂÌ˲
ã‡ÏÔ‡ Bluephase N – ˝ÚÓ Ò‚ÂÚÓ‰ËÓ‰Ì˚È ÔË·Ó Ò
ËÁÎÛ˜ÂÌËÂÏ ÒËÌÂ„Ó ÒÔÂÍÚ‡ ‚˚ÒÓÍÓÈ ÏÓ˘ÌÓÒÚË ‰Îfl
ÔÓÎËÏÂËÁ‡ˆËË Ò‚ÂÚÓÓÚ‚Âʉ‡ÂÏ˚ı
ÒÚÓχÚÓÎӄ˘ÂÒÍËı χÚÂˇÎÓ‚ ÌÂÔÓÒ‰ÒÚ‚ÂÌÌÓ Û
ÒÚÓχÚÓÎӄ˘ÂÒÍÓÈ ÛÒÚ‡ÌÓ‚ÍË. ä ËÒÔÓθÁÓ‚‡Ì˲ ÔÓ
̇Á̇˜ÂÌ˲ ÓÚÌÓÒËÚÒfl Òӷβ‰ÂÌË ÂÍÓÏẨ‡ˆËÈ
̇ÒÚÓfl˘ÂÈ ËÌÒÚÛ͈ËË.
èÓ͇Á‡ÌËfl
Å·„Ó‰‡fl ¯ËÓÍÓÔÓÎÓÒÌÓÏÛ ÔÓÎË‚ÓÎÌÓ‚ÓÏÛ ÒÔÂÍÚÛ
·ÏÔÓÈ Bluephase N ÏÓÊÌÓ ÔÓ‚Ó‰ËÚ¸ ÔÓÎËÏÂËÁ‡-
ˆË˛ ‚ÒÂı Ò‚ÂÚÓÓÚ‚Âʉ‡ÂÏ˚ı ÒÚÓχÚÓÎӄ˘ÂÒÍËı
χÚÂˇÎÓ‚ ‚ ‰Ë‡Ô‡ÁÓÌ ‰ÎËÌ˚ Ò‚ÂÚÓ‚ÓÈ ‚ÓÎÌ˚
385–515 ÌÏ. ä Ú‡ÍËÏ Ï‡ÚÂË‡Î‡Ï ÓÚÌÓÒflÚÒfl ÔÎÓÏ·Ë-
Ó‚Ó˜Ì˚ χÚÂˇÎ˚, ·Ó̉ËÌ„Ë/‡‰„ÂÁË‚˚, ÔӉͷ‰Ó˜-
Ì˚ χÚÂˇÎ˚, ·ÈÌÂ˚, χÚÂˇÎ˚ ‰Îfl
Á‡Ô˜‡Ú˚‚‡ÌËfl ÙËÒÒÛ,
χÚÂˇÎ˚ ‰Îfl ‚ÂÏÂÌÌ˚ı ÂÒÚ‡‚‡ˆËÈ, ‡ Ú‡ÍÊÂ
ÙËÍÒËÛ˛˘Ë ÍÓÏÔÓÁËÚ˚ ‰Îfl ·ÂÍÂÚÓ‚ Ë
ÒÚÓχÚÓÎӄ˘ÂÒÍËı ÍÓÌÒÚÛ͈ËÈ, ͇Í, ̇ÔËÏÂ,
ÍÂ‡Ï˘ÂÒÍË ‚Í·‰ÍË ÚËÔ‡ Inlay.
ê‡Á˙flÒÌÂÌË Á̇ÍÓ‚
ç‰ÓÔÛÒÚËÏÓ ÔËÏÂÌÂÌËÂ
ëËÏ‚ÓÎ˚ ̇ ÔË·ÓÂ:
Ñ‚ÓÈ̇fl ËÁÓÎflˆËfl
(ÄÔÔ‡‡Ú Í·ÒÒ‡ Á‡˘ËÚ˚ II)
ᇢËÚ‡ ÓÚ Û‰‡‡ ˝ÎÂÍÚÓÚÓÍÓÏ
(ÚËÔ ‡ÔÔ‡‡Ú‡ BF)
ëӷβ‰‡Ú¸ Ú·ӂ‡ÌËfl ËÌÒÚÛ͈ËË
ëӷβ‰‡Ú¸ Ú·ӂ‡ÌËfl ËÌÒÚÛ͈ËË
éÒÚÓÓÊÌÓ!
èË·Ó ÌÂθÁfl ÛÚËÎËÁËÓ‚‡Ú¸ Í‡Í Ó·˚˜Ì˚È
·˚ÚÓ‚ÓÈ ÏÛÒÓ. àÌÙÓχˆË˛ ÔÓ
ÛÚËÎËÁ‡ˆËË ÔË·Ó‡ ‚˚ ̇ȉÂÚ ̇
ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘ÂÈ ‰Óχ¯ÌÂÈ ÒÚ‡ÌˈÂ
̇ˆËÓ̇θÌÓ„Ó ‚·-Ò‡ÈÚ‡ Ivoclar Vivadent.
àÒÔÓθÁÓ‚‡ÌË ÚÓθÍÓ ‚ Á‡˘Ë˘ÂÌÌÓÏ
ÅÂÁÓÔ‡ÒÌÓÒÚ¸
ã‡ÏÔ‡ Bluephase N –˝ÚÓ ˝ÎÂÍÚ˘ÂÒÍËÈ ÔË·ÓË
fl‚ÎflÂÚÒfl ωˈËÌÒÍËÏ ÔÓ‰ÛÍÚÓÏ, ÍÓÚÓ˚È ÔÓ‰ÎÂÊËÚ
ÌÓÏ‡Ï IEC 60601-1 (EN 60601-1) Ë EMV, ‡ Ú‡ÍÊÂ
‰ÂÈÒÚ‚Û˛˘ËÏ ÌÓÏ‡Ï ÔÓ Ï‰ˈËÌÒÍËÏ ÚÓ‚‡‡Ï
93/42/EWG. èË·Ó ÒÓÓÚ‚ÂÚÒÚ‚ÛÂÚ ‰ÂÈÒÚ‚Û˛˘ËÏ
ÌÓÏ‡Ï Öë Ë Í·ÒÒËÙˈËÛÂÚÒfl Í‡Í Ò‚ÂÚÓ‰ËÓ‰Ì˚È
ÔË·Ó Í·ÒÒ‡ 2.
èË·Ó ·˚Î ‚˚ÔÛ˘ÂÌ Á‡‚Ó‰ÓÏ ‚ ̇‰ÂÊÌÓÏ Ë ÚÂıÌ˘ÂÒÍË
·ÂÁÛÔ˜ÌÓÏ ÒÓÒÚÓflÌËË. óÚÓ·˚ ÒÓı‡ÌflÚ¸ Ë
ÔÓ‰‰ÂÊË‚‡Ú¸ ˝ÚÓ ÒÓÒÚÓflÌËÂ, ‡ Ú‡ÍÊ ӷÂÒÔ˜˂‡Ú¸
·ÂÁÓÔ‡ÒÌÛ˛ ‡·ÓÚÛ Ò ÔË·ÓÓÏ, ÒΉÛÂÚ Òӷβ‰‡Ú¸
ÂÍÓÏẨ‡ˆËË ‰‡ÌÌÓÈ ËÌÒÚÛ͈ËË. óÚÓ·˚ ËÁ·Âʇڸ
Ô˘ËÌÂÌËfl ‚‰‡ Ô‡ˆËÂÌÚ‡Ï, ÔÓθÁÓ‚‡ÚÂÎflÏ Ë
ÚÂÚ¸ËÏ Îˈ‡Ï, ÓÒÓ·ÂÌÌÓ ÒΉÛÂÚ Ó·‡ÚËÚ¸ ‚ÌËχÌËÂ
̇ ÒÎÂ‰Û˛˘Ë ÔÓÎÓÊÂÌËfl:
å‡ÚÂˇÎ˚, ÔÓÎËÏÂËÁ‡ˆËfl ÍÓÚÓ˚ı ‡ÍÚË-
‚ËÛÂÚÒfl Ì ‚ ‰Ë‡Ô‡ÁÓÌ ҂ÂÚÓ‚ÓÈ ‚ÓÎÌ˚
385–515 ÌÏ (‚ ̇ÒÚÓfl˘ËÈ ÏÓÏÂÌÚ Ú‡ÍËÂ
χÚÂˇÎ˚ Ì ËÁ‚ÂÒÚÌ˚). Ç ÒÓÏÌËÚÂθÌ˚ı
ÒÎÛ˜‡flı, ÍÓ„‰‡ Ç˚ Ì ۂÂÂÌ˚ ‚ ÔÓ‰ÛÍÚÂ,
Ï˚ ÂÍÓÏẨÛÂÏ Ç‡Ï Ó·‡ÚËÚ¸Òfl Í
ÔÓËÁ‚Ó‰ËÚÂβ χÚÂˇ·.
èË·Ó ÌÂθÁfl ËÒÔÓθÁÓ‚‡Ú¸ Ë Áaflʇڸ
‚·ÎËÁË ‚ÓÒÔ·ÏÂÌfl˛˘ËıÒfl ËÎË ‚Á˚‚˜aÚ˚ı
‚¢ÂÒÚ‚.
èÂÂÌÓÒÌ˚Â Ë ÏÓ·ËθÌ˚ ‚˚ÒÓÍÓ˜‡ÒÚÓÚÌ˚Â
ÍÓÏÏÛÌË͇ˆËÓÌÌ˚ Ò‰ÒÚ‚‡ ÏÓ„ÛÚ ‚ÎËflÚ¸
̇ ‡·ÓÚÛ Ï‰ˈËÌÒÍÓÈ ÚÂıÌËÍË. èÓ˝ÚÓÏÛ
Ó‰ÌÓ‚ÂÏÂÌÌÓ ËÒÔÓθÁÓ‚‡ÌË ÏÓ·ËθÌ˚ı
ÚÂÎÂÙÓÌÓ‚ Ë ‡ÔÔ‡‡Ú‡ Ì ‰ÓÔÛÒÚËÏÓ.
éÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚ¸ ÔÓθÁÓ‚‡ÚÂÎfl Ë ÔÓËÁ‚Ó‰ËÚÂÎfl
ã‡ÏÔ‡ Bluephase N ÏÓÊÂÚ ·˚Ú¸ ËÒÔÓθÁÓ‚‡Ì‡
ËÒÍβ˜ËÚÂθÌÓ ÚÓθÍÓ ÔÓ Ì‡Á̇˜ÂÌ˲. ã˛·ÓÂ
‰Û„Ó ÔËÏÂÌÂÌË ‚ ËÌ˚ı ˆÂÎflı fl‚ÎflÂÚÒfl
ËÒÔÓθÁÓ‚‡ÌËÂÏ Ì ÔÓ Ì‡Á̇˜ÂÌ˲. èÓËÁ‚Ó‰ËÚÂθ
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 19
77
Ì ÌÂÒÂÚ ÓÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚË Á‡ ‚‰, Ô˘ËÌÂÌÌ˚È
ÂÁÛθڇÚ ËÒÔÓθÁÓ‚‡ÌËfl ÔË·Ó‡ Ì ÔÓ Ì‡Á̇˜ÂÌ˲
ËÎË ÌÂÍÓÂÍÚÌÓ„Ó ÛıÓ‰‡ ÁÌËÏ.
äÓÏ ÚÓ„Ó, ÔÓÚ·ËÚÂθ Ó·flÁ‡Ì ÔÓ‰ Ò‚Ó˛
ÓÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚ¸ ÔÓ‚ÂËÚ¸ ·ÏÔÛ Bluephase N ÔÂ‰
 ËÒÔÓθÁÓ‚‡ÌËÂÏ Ì‡ ÒÓÓÚ‚ÂÚÒÚ‚ËÂ Ë ‚ÓÁÏÓÊÌÓÒÚ¸
ÔËÏÂÌÂÌËfl ‰Îfl ÔÓÒÚ‡‚ÎÂÌÌ˚ı ˆÂÎÂÈ. éÒÓ·ÂÌÌÓ ˝ÚÓ
͇҇ÂÚÒfl ÚÂı ÒÎÛ˜‡Â‚, ÍÓ„‰‡ ‚ ÌÂÔÓÒ‰ÒÚ‚ÂÌÌÓÈ ·ÎË-
ÁÓÒÚË ÓÚ ÔË·Ó‡ Ë Ó‰ÌÓ‚ÂÏÂÌÌÓ Ò ÌËÏ ‡·ÓÚ‡˛Ú Ò
‰Û„ËÏË ‡ÔÔ‡‡Ú‡ÏË.
ÑÓÔÛÒ͇ÂÚÒfl ËÒÔÓθÁÓ‚‡Ú¸ ÚÓθÍÓ ÓË„Ë̇θÌ˚Â
Á‡Ô‡ÒÌ˚ ˜‡ÒÚË Ë ÔË̇‰ÎÂÊÌÓÒÚË ÙËÏ˚ Ivoclar
Vivadent (ÒÏ. ‡Á‰ÂÎ èË̇‰ÎÂÊÌÓÒÚË). èÓËÁ‚Ó‰Ë-
ÚÂθ Ì ÌÂÒÂÚ ÓÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚË Á‡ ‚‰, Ô˘ËÌÂÌÌ˚È
‚ ÂÁÛθڇÚ ËÒÔÓθÁÓ‚‡ÌËfl ËÌ˚ı Á‡Ô‡ÒÌ˚ı ˜‡ÒÚÂÈ
ËÎË ÔË̇‰ÎÂÊÌÓÒÚÂÈ.
ë‚ÂÚÓ‚Ó‰ – ˝ÚÓ ‡·Ó˜‡fl ˜‡ÒÚ¸ Ë ‚Ó ‚ÂÏfl ‡·ÓÚ˚ ‚
ÏÂÒÚ ÒÓ‰ËÌÂÌËfl Ò Ì‡ÍÓ̘ÌË-ÍÓÏ ÏÓÊÂÚ
̇„‚‡Ú¸Òfl ‰Ó χÍÒ. 45°C.
ꇷӘ ̇ÔflÊÂÌËÂ
èÂ‰ ‚Íβ˜ÂÌËÂÏ Î‡ÏÔ˚ ÒΉÛÂÚ Û‰ÓÒÚÓ‚ÂËÚ¸Òfl, ˜ÚÓ
a) ̇ÔflÊÂÌËÂ, Û͇Á‡ÌÌÓ ̇ Ú‡·Î˘ÍÂ, ÒÓÓÚ‚ÂÚÒÚ‚ÛÂÚ
̇ÔflÊÂÌ˲ ‚ ÒÂÚË Ë
·) ÔË·Ó ËÏÂÂÚ ÍÓÏ̇ÚÌÛ˛ ÚÂÏÔÂ‡ÚÛÛ.
èË ‡·ÓÚ ÓÚ‰ÂθÌÓ Ò ‡ÍÍÛÏÛÎflÚÓÓÏ ËÎË ÒÂÚ‚˚Ï
·ÎÓÍÓÏ – ̇ÔËÏÂ, ÔË ‚‚Ӊ ‚ ˝ÍÒÔÎÛ‡Ú‡ˆË˛ ËÎË
ÔË ‡·ÓÚ ˜ÂÂÁ ͇·Âθ Click & Cure, – ËÁ·Â„‡Ú¸
ÍÓÌÚ‡ÍÚ‡ Ò Ô‡ˆËÂÌÚÓÏ ËÎË ÚÂÚ¸ËÏË Îˈ‡ÏË. ç ‰ÓÚ‡-
„Ë‚‡Ú¸Òfl ‰Ó Ó„ÓÎÂÌÌ˚ı ˝ÎÂÍÚ˘ÂÒÍËı ÍÓÌÚ‡ÍÚÓ‚ ‡ÍÍÛ-
ÏÛÎflÚÓ‡ ËÎË ÒÓ‰ËÌËÚÂθÌÓ„Ó ¯ÚÂÍÂ‡ (ÒÂÚ‚ӄÓ
·ÎÓ͇).
åÂ˚ Ô‰ÓÒÚÓÓÊÌÓÒÚË
ÖÒÎË ‚ÓÁÌËÍÎÓ Ô‰ÔÓÎÓÊÂÌËÂ, ˜ÚÓ ·ÂÁÓÔ‡ÒÌÓÂ
ËÒÔÓθÁÓ‚‡ÌË ÔË·Ó‡ Ì‚ÓÁÏÓÊÌÓ, ÔË·Ó ÒΉÛÂÚ
ÓÚÍβ˜ËÚ¸ ÓÚ ‡ÍÍÛÏÛÎflÚÓ‡ Ë ˝ÎÂÍÚÓÒÂÚË Ë ÔËÌflÚ¸
ÏÂ˚, ˜ÚÓ·˚ ÓÌ Ì ·˚Î ‚Íβ˜ÂÌ ‚ ÂÁÛθڇÚ ÌÂÓÒ‚Â-
‰ÓÏÎÂÌÌÓÒÚË ÔÂÒÓ̇·. í‡ÍÓ Ô‰ÔÓÎÓÊÂÌË ÏÓÊÂÚ
‚ÓÁÌËÍÌÛÚ¸, ̇ÔËÏÂ, ÔË ‚ˉËÏ˚ı ÔÓ‚ÂʉÂÌËflı
ÔË·Ó‡ ËÎË Ó„‡Ì˘ÂÌÌÓÏ ‚ÓÒÔÓËÁ‚‰ÂÌËË Â„Ó
ÙÛÌ͈ËÈ. èÓÎÌÓ ÓÚÍβ˜ÂÌË ÔË·Ó‡ ÓÚ ÒÂÚË „‡‡Ì-
ÚËÛÂÚÒfl ÚÓθÍÓ ÔË ÓÚÒÓ‰ËÌÂÌËË ÒÂÚÂ‚Ó„Ó Í‡·ÂÎfl ÓÚ
ÓÁÂÚÍË.
ᇢËÚ‡ „·Á
ëΉÛÂÚ ËÁ·Â„‡Ú¸ ÔflÏÓ„Ó ËÎË ÌÂÔflÏÓ„Ó ÔÓÔ‡‰‡ÌËfl
ËÁÎÛ˜ÂÌËfl ‚ „·Á‡. ÑÎËÚÂθÌÓ ÓÒ‚Â˘ÂÌË ÌÂÔËflÚÌÓ
‰Îfl „·Á Ë ÏÓÊÂÚ Ì‡ÌÂÒÚË ËÏ ‚‰. èÓ˝ÚÓÏÛ ÂÍÓÏÂÌ-
‰ÛÂÚÒfl ËÒÔÓθÁÓ‚‡Ú¸ Á‡˘ËÚÌ˚È ˝Í‡Ì, ÔÓÒÚ‡‚ÎflÂÏ˚È
‚ÏÂÒÚÂ Ò ‡ÔÔ‡‡ÚÓÏ. ãˈ‡, ÍÓÚÓ˚ ‚ÓÓ·˘Â ˜Û‚ÒÚ‚Ë-
ÚÂθÌ˚ Í Ò‚ÂÚÛ, ÔËÌËχ˛Ú ωË͇ÏÂÌÚ˚ ÔÓÚË‚ Ò‚ÂÚÓ-
˜Û‚ÒÚ‚ËÚÂθÌÓÒÚË ËÎË ÙÓÚÓ˜Û‚ÒÚ‚ËÚÂθÌ˚Â
ωË͇ÏÂÌÚ˚, ÔÂÂÌÂүˠÓÔÂ‡ˆË˛ ̇ „·Á‡, ËÎË
‡·ÓÚ‡˛˘Ë ‚ Ú˜ÂÌË ‰ÎËÚÂθÌÓ„Ó ‚ÂÏÂÌË Ò ˝ÚËÏ
‡ÔÔ‡‡ÚÓÏ ËÎË ‚ ÌÂÔÓÒ‰ÒÚ‚ÂÌÌÓÈ ·ÎËÁÓÒÚË ÓÚ Ì„Ó,
Ì ‰ÓÎÊÌ˚ ÔÓ‰‚Â„‡Ú¸Òfl ‰ÂÈÒڂ˲ ˝ÚÓ„Ó ‡ÔÔ‡‡Ú‡,
‰ÓÎÊÌ˚ ÌÓÒËÚ¸ Á‡˘ËÚÌ˚ (Ó‡ÌÊ‚˚Â) Ó˜ÍË, ÍÓÚÓ˚Â
‡·ÒÓ·ËÛ˛Ú Ò‚ÂÚ Ò ‰ÎËÌÓÈ ‚ÓÎÌ˚ ‰Ó 515 ÌÏ.
ÄÍÍÛÏÛÎflÚÓ
ÇÌËχÌËÂ: ËÒÔÓθÁÓ‚‡Ú¸ ÚÓθÍÓ ÓË„Ë̇θÌ˚ ‰ÂÚ‡ÎË
– ÓÒÓ·ÂÌÌÓ ˝ÚÓ Í‡Ò‡ÂÚÒfl ‡ÍÍÛÏÛÎflÚÓ‡ Ë Á‡fl‰ÌÓ„Ó
·ÎÓ͇ ÔÓËÁ‚Ó‰ÒÚ‚‡ Ivoclar Vivadent. ÄÍÍÛÏÛÎflÚÓ ÌÂ
Á‡ÍÓ‡˜Ë‚‡Ú¸. ç ı‡ÌËÚ¸ Â„Ó ÔË ÚÂÏÔÂ‡ÚÛ ‚˚¯Â
40 °C (ËÎË Í‡ÚÍÓ‚ÂÏÂÌÌÓ ÔË 60°C). ï‡ÌËÚ¸ ÚÓθÍÓ
‚ Á‡flÊÂÌÌÓÏ ÒÓÒÚÓflÌËË. ÑÎËÚÂθÌÓÒÚ¸ ı‡ÌÂÌËfl ÌÂ
‰ÓÎÊ̇ Ô‚˚¯‡Ú¸ 6 ÏÂÒflˆÂ‚. ÇÁ˚‚ÓÓÔ‡ÒÂÌ ÔË
ÛÚËÎËÁ‡ˆËË ‚ ÓÚÍ˚ÚÓÏ Ô·ÏÂÌË.
é·‡ÚËÚ ‚ÌËχÌËÂ, ˜ÚÓ ÂÒÎË ÎËÚËÈ-ÔÓÎËÏÂÌ˚È ‡ÍÍÛ-
ÏÛÎflÚÓ ËÒÔÓθÁÛÂÚÒfl Ì ‚ ÒÓÓÚ‚ÂÚÒÚ‚ËË Ò Ì‡Á̇˜ÂÌË-
ÂÏ Ë ËÌÒÚÛ͈ËÂÈ ËÎË ËÏÂÂÚ ÏÂı‡Ì˘ÂÒÍËÂ
ÔÓ‚ÂʉÂÌËfl, ÂÒÚ¸ ‚ÂÓflÚÌÓÒÚ¸ ‚Á˚‚‡, ‚ÓÒÔ·ÏÂÌÂÌËfl
ËÎË ÔÓfl‚ÎÂÌËfl ‰˚χ. èÓ‚ÂʉÂÌÌ˚ ÎËÚËÈ-ÔÓÎËÏÂ-
Ì˚ ‡ÍÍÛÏÛÎflÚÓ˚ ËÒÔÓθÁÓ‚‡Ú¸ ÌÂθÁfl.
ùÎÂÍÚÓÎËÚ˚, ˝ÎÂÍÚÓÎËÚ˘ÂÒÍË ԇ˚, Ó·‡ÁÛ˛˘ËÂÒfl
ÔË ‚Á˚‚Â, ‚ÓÒÔ·ÏÂÌÂÌËË ËÎË Á‡‰˚ÏÎÂÌËË, Ó͇Á˚-
‚‡˛Ú ÚÓÍÒ˘ÂÒÍÓÂ Ë ‡Á˙‰‡˛˘Â ‰ÂÈÒÚ‚ËÂ. èË
ÔÓÔ‡‰‡ÌËË ‚ „·Á‡ Ë Ì‡ ÍÓÊÛ Ò‡ÁÛ Ê ÔÓÏ˚Ú¸ ·Óθ-
¯ËÏ ÍÓ΢ÂÒÚ‚ÓÏ ‚Ó‰˚. àÁ·Â„‡ÈÚ ‚‰˚ı‡ÌËfl Ô‡Ó‚.
èË ÛıÛ‰¯ÂÌËË Ò‡ÏÓ˜Û‚ÒÚ‚Ëfl Ó·‡ÚËÚÂÒ¸ Í ‚‡˜Û.
Ç˚‰ÂÎÂÌË ÚÂÔ·
Ç ‡·ÓÚ β·Ó„Ó ÏÓ˘ÌÓ„Ó ÔÓÎËÏÂËÁ‡ˆËÓÌÌÓ„Ó ÔË-
·Ó‡ ËÁÎÛ˜ÂÌË ҂ÂÚ‡ ‚˚ÒÓÍÓÈ ËÌÚÂÌÒË‚ÌÓÒÚË ‚Ò„‰‡
ÒÓÔÓ‚Óʉ‡ÂÚÒfl ‚˚‰ÂÎÂÌËÂÏ ÚÂÔ·. èË ‰ÎËÚÂθÌÓÏ
ÓÒ‚Â˘ÂÌËË ÔÛθÔ˚ ËÎË Ïfl„ÍËı Ú͇ÌÂÈ ËÏ ÏÓÊÂÚ ·˚Ú¸
̇ÌÂÒÂÌ ‚‰ Ò ÌÂÓ·‡ÚËÏ˚ÏË ËÎË Ó·‡ÚËÏ˚ÏË
ÔÓÒΉÒÚ‚ËflÏË. èÓ˝ÚÓÏÛ ÒΉÛÂÚ Òӷβ‰‡Ú¸ Ô‰ÔË-
Ò‡ÌÌÓ ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË. çÂÔÂÏÂÌÌÓ ÒΉÛÂÚ
ËÁ·Â„‡Ú¸ ÌÂÔÂ˚‚ÌÓ„Ó ÓÒ‚Â˘ÂÌËfl ‚ Ú˜ÂÌË ·ÓÎÂÂ
˜ÂÏ 20 ÒÂÍÛ̉ Ó‰ÌÓÈ Ë ÚÓÈ Ê ÔÓ‚ÂıÌÓÒÚË ÁÛ·‡, ‡
Ú‡ÍÊ ÔflÏÓ„Ó ÍÓÌÚ‡ÍÚ‡ Ò ‰ÂÒÌÓÈ, ÒÎËÁËÒÚÓÈ
Ó·ÓÎÓ˜ÍÓÈ ËÎË ÍÓÊÂÈ. Ç ÒÎÛ˜‡Â ‚˚ÔÓÎÌÂÌËfl
ÂÒÚ‡‚‡ˆËË ÌÂÔflÏ˚Ï ÏÂÚÓ‰ÓÏ ÒΉÛÂÚ ‡·ÓÚ‡Ú¸ Ò
ÔÂÂ˚‚‡ÏË, ‚ÂÏÂÌÌÓÈ ËÌÚÂ‚‡Î ‰ÓÎÊÂÌ ÒÓÒÚ‡‚ÎflÚ¸
20 ÒÂÍÛ̉, ËÎË ÔË Ì‡ÛÊÌÓÏ Óı·ʉÂÌËË Ò ÔÓÏÓ˘¸˛
ÒÚÛË ‚ÓÁ‰Ûı‡.
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 20
78
èÓ‚Â¸Ú ÍÓÏÔÎÂÍÚÌÓÒÚ¸ ÔÓÒÚ‡‚ÍË (ÒÏ. é·˙ÂÏ ÔÓÒÚ‡‚-
ÍË) Ë Ì‡Î˘Ë Ú‡ÌÒÔÓÚÌ˚ı ÔÓ‚ÂʉÂÌËÈ. Ç ÒÎÛ˜‡Â,
ÂÒÎË Í‡ÍËÂ-ÎË·Ó ˜‡ÒÚË ÔÓ‚ÂʉÂÌ˚ ËÎË ÓÚÒÛÚÒÚ‚Û˛Ú,
ÌÂωÎÂÌÌÓ Ò‚flÊËÚÂÒ¸ Ò ÔÓ‰‡‚ˆÓÏ ËÎË ÒÂ‚ËÒÌÓÈ
ÒÎÛÊ·ÓÈ.
á‡fl‰Ì˚È ·ÎÓÍ
èÂ‰ ‚Íβ˜ÂÌËÂÏ ÒΉÛÂÚ Û‰ÓÒÚÓ‚ÂËÚ¸Òfl, ˜ÚÓ Û͇-
Á‡ÌÌÓ ̇ Ú‡·Î˘Í ̇ÔflÊÂÌË ÒÓÓÚ‚ÂÚÒÚ‚ÛÂÚ Ì‡Ôfl-
ÊÂÌ˲ ‚ ÒÂÚË. 퇷΢͇ ̇ıÓ‰ËÚÒfl ̇ ÌËÊÌÂÈ ˜‡ÒÚË
Á‡fl‰ÌÓ„Ó ·ÎÓ͇.
èÓÒÚ‡‚¸Ú Á‡fl‰Ì˚È ·ÎÓÍ Ì‡ ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘Û˛ Ó‚ÌÛ˛
ÔÓ‚ÂıÌÓÒÚ¸ ÒÚÓ·.
쉇ÎËÚ Á‡˘ËÚÌÛ˛ ÔÎÂÌÍÛ Ò ÒÓ‰ËÌËÚÂθÌÓ„Ó ¯ÚÂÍÂ‡
ÒÂÚÂ‚Ó„Ó ·ÎÓ͇. ëÓ‰ËÌËÚÂθÌ˚È ¯ÚÂÍÂ ÒÂÚÂ‚Ó„Ó ·ÎÓ-
͇ ‚ÒÚ‡‚ËÚ¸ ÔÓ‰ Û„ÎÓÏ ‚ Á‡fl‰Ì˚È ·ÎÓÍ, ‡ Á‡ÚÂÏ Ò Î„-
ÍËÏ ‰‡‚ÎÂÌË Á‡ÙËÍÒËÓ‚‡Ú¸ (Ú‡Í ˜ÚÓ·˚ ˜Û‚ÒÚ‚Ó‚‡ÎÒfl
Ë ·˚Î ÒÎ˚¯ÂÌ ˘ÂΘÓÍ).
èÓ‰ÒÓ‰ËÌËÚ¸ ÒÂÚ‚ÓÈ Í‡·Âθ Í ÒÂÚË Ë ÒÂÚ‚ÓÏÛ ·ÎÓÍÛ.
é·ÓÁ̇˜ÂÌË Power ON c ΂ÓÈ ÒÚÓÓÌ˚ ÍÓÔÛÒ‡ Á‡„Ó-
ËÚÒfl ÒËÌËÏ ˆ‚ÂÚÓÏ (ÒÏ. ‡Á‰ÂÎ é·ÓÁ̇˜ÂÌËfl ̇
Á‡fl‰ÌÓÏ ·ÎÓÍÂ).
ç‡ÍÓ̘ÌËÍ
èÂ‰ ÏÓÌÚ‡ÊÓÏ Ò‚ÂÚÓ‚Ó‰‡ ÒΉÛÂÚ Û‰‡ÎËÚ¸ Á‡˘ËÚÌÛ˛
ÔÎÂÌÍÛ Ò ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘Â„Ó ÓÚ‚ÂÒÚËfl ̇ ̇ÍÓ̘ÌËÍÂ.
ë‚ÂÚÓ‚Ó‰ ‚ÒÚ‡‚ËÚ¸ ‚ ̇ÍÓ̘ÌËÍ Î„ÍËÏ ‚‡˘ÂÌËÂÏ.
á‡ÚÂÏ ÛÒÚ‡ÌÓ‚ËÚ¸ ̇ Ò‚ÂÚÓ‚Ó‰ Á‡˘ËÚÌ˚È ˝Í‡Ì.
Ç‚Ó‰ ‚ ˝ÍÒÔÎÛ‡Ú‡ˆË˛
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 21
79
ÄÍÍÛÏÛÎflÚÓ
èÂ‰ ‚‚Ó‰ÓÏ ‚ ˝ÍÒÔÎÛ‡Ú‡ˆË˛ ‡ÍÍÛÏÛÎflÚÓ ‰ÓÎÊÂÌ
·˚Ú¸ ÔÓÎÌÓÒÚ¸˛ Á‡flÊÂÌ!
èÓÎÌÓÒÚ¸˛ Á‡flÊÂÌÌÓ„Ó ‡ÍÍÛÏÛÎflÚÓ‡ ı‚‡Ú‡ÂÚ
ÔËÏÂÌÓ Ì‡ 60 ÏËÌÛÚ ÔÓÎËÏÂËÁ‡ˆËË.
ÄÍÍÛÏÛÎflÚÓ ‚ÒÚ‡‚ËÚ¸ ‚ ̇ÍÓ̘ÌËÍ ÔÓ ÔflÏÓÈ, Ú‡Í,
˜ÚÓ·˚ ˜Û‚ÒÚ‚Ó‚‡ÎÒfl Ë ÒÎ˚¯‡ÎÒfl ˘ÂΘÓÍ.
ç ÔËÏÂÌflfl ÛÒËÎËÂ, ‚ÒÚ‡‚ËÚ¸ ̇ÍÓ̘ÌËÍ ‚ ÓÚ‚ÂÒÚËÂ
̇ Á‡fl‰ÌÓÏ ·ÎÓÍÂ. ÖÒÎË ÔËÏÂÌflÂÚÒfl „Ë„ËÂÌ˘ÂÒ͇fl
Á‡˘ËÚ‡,  ÒΉÛÂÚ Û‰‡ÎËÚ¸ ÔÂ‰ Á‡fl‰ÍÓÈ
‡ÍÍÛÏÛÎflÚÓ‡.
èÓ ‚ÓÁÏÓÊÌÓÒÚË ÔË·Ó ÔÓÒÚÓflÌÌÓ ËÒÔÓθÁÓ‚‡Ú¸ Ò
ÔÓÎÌÓÒÚ¸˛ Á‡flÊÂÌÌ˚Ï ‡ÍÍÛÏÛÎflÚÓÓÏ – ˝ÚÓ Ó·ÂÒ-
Ô˜˂‡ÂÚ ‰ÎËÚÂθÌ˚È ÒÓÍ ÒÎÛÊ·˚. èÓ˝ÚÓÏÛ
ÂÍÓÏẨÛÂÚÒfl ÔÓÒÎÂ Í‡Ê‰Ó„Ó Ô‡ˆËÂÌÚ‡ ‚ÒÚ‡‚ÎflÚ¸
̇ÍÓ̘ÌËÍ ‚ Á‡fl‰Ì˚È ·ÎÓÍ. á‡fl‰Í‡ ÔÓÎÌÓÒÚ¸˛
‡Áfl‰Ë‚¯Â„ÓÒfl ‡ÍÍÛÏÛÎflÚÓ‡ ‰ÎËÚÒfl 2 ˜‡Ò‡.
ìÓ‚Â̸ Á‡fl‰ÍË ‡ÍÍÛÏÛÎflÚÓ‡
ìÓ‚Â̸ Á‡fl‰ÍË ‡ÍÍÛÏÛÎflÚÓ‡ ÓÚÓ·‡Ê‡ÂÚÒfl ̇
‰ËÒÔΠÔË ‚Íβ˜ÂÌÌÓÏ Ì‡ÍÓ̘ÌËÍ ÒÎÂ‰Û˛˘ËÏ
Ó·‡ÁÓÏ:
èÓÎÌÓÒÚ¸˛ Á‡flÊÂÌÌ˚È ‡ÍÍÛÏÛÎflÚÓ
(‡·Ó˜Â ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË
ÒÓÒÚ‡‚ÎflÂÚ ÔËÏÂÌÓ 60 ÏËÌÛÚ).
ç‡ÔÓÎÓ‚ËÌÛ Á‡flÊÂÌÌ˚È ‡ÍÍÛÏÛÎflÚÓ
êÂÁÂ‚ (Ç ÂÊËÏ êÂÁÂ‚ ÔÓÒΉÌflfl
·‡Î͇ ̇ ‰Ë‡„‡ÏÏ Á‡fl‰ÍË ÒÚ‡ÌÓ‚ËÚÒfl
Í‡ÒÌÓÈ. ÄÍÍÛÏÛÎflÚÓ ÒΉÛÂÚ Á‡fl‰ËÚ¸
Í‡Í ÏÓÊÌÓ ÒÍÓÂÂ).
ÖÒÎË ‡ÍÍÛÏÛÎflÚÓ Á‡flʇÂÚÒfl, ÔÂʉÂ
˜ÂÏ ‰ËÒÔÎÂÈ ÔÂÂȉÂÚ ‚ ÂÊËÏ Stand-
by, ̇ ‰ËÒÔΠ‰ÓÎÊ̇ ÔÓfl‚ËÚ¸Òfl Í‡ÚÍÓ-
‚ÂÏÂÌ̇fl ̇‰ÔËÒ¸ “Charging Battery“.
èË ÔÓÎÌÓÒÚ¸˛ ‡ÁflÊÂÌÌÓÏ
‡ÍÍÛÏÛÎflÚÓ ̇ÍÓ̘ÌËÍ ‡‚ÚÓχÚ˘ÂÒÍË
ÔÂÂÍβ˜‡ÂÚÒfl ‚ ÒÓÒÚÓflÌË Stand-by.
àÁÎÛ˜ÂÌË ·Óθ¯Â Ì ‡ÍÚË‚ËÛÂÚÒfl,
ÔÓ„‡ÏÏÛ Ë ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË
Á‡‰‡Ú¸ Ì‚ÓÁÏÓÊÌÓ. é‰Ì‡ÍÓ, ‚ Ú‡ÍËı
ÒÎÛ˜‡flı ̇ÍÓ̘ÌËÍÓÏ ÏÓÊÌÓ ‡·ÓÚ‡Ú¸,
ËÒÔÓθÁÛfl ÙÛÌÍˆË˛ ‡·ÓÚ˚ ˜ÂÂÁ
͇·Âθ Click & Cure.
ÄÍÍÛÏÛÎflÚÓ fl‚ÎflÂÚÒfl ·˚ÒÚÓ ËÁ̇¯Ë‚‡˛˘ÂÈÒfl
˜‡ÒÚ¸˛, Ó·˚˜ÌÓ ˜ÂÂÁ 2 Ò ÔÓÎÓ‚ËÌÓÈ „Ó‰‡ ÓÌ ÔÓ‰ÎÂÊËÚ
Á‡ÏÂÌÂ. «ÇÓÁ‡ÒÚ» ‡ÍÍÛÏÛÎflÚÓ‡ ÏÓÊÌÓ ÛÁ̇ڸ ÔÓ
̇ÍÎÂÈÍ ̇ ÌÂÏ.
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 22
80
ꇷÓÚ‡ ˜ÂÂÁ ͇·Âθ Click & Cure
ã‡ÏÔÓÈ Bluephase N ‚ β·Ó ‚ÂÏfl ÏÓÊÌÓ ‡·ÓÚ‡Ú¸
˜ÂÂÁ ͇·Âθ, ‡ ÓÒÓ·ÂÌÌÓ ÔË ÔÓÎÌÓÒÚ¸˛ ‡ÁflÊÂÌÌÓÏ
‡ÍÍÛÏÛÎflÚÓÂ.
ÑÎfl ˝ÚÓ„Ó ÌÂÓ·ıÓ‰ËÏÓ ‚˚ÌÛÚ¸ ‡ÍÍÛÏÛÎflÚÓ ËÁ ̇ÍÓ̘-
ÌË͇. èÓÒΠ˝ÚÓ„Ó ÓÚÒÓ‰ËÌËÚ¸ ¯ÚÂÍÂ ÒÂÚÂ‚Ó„Ó ·ÎÓ͇
ÔËÚ‡ÌËfl Ò ÌËÊÌÂÈ ÒÚÓÓÌ˚ Á‡fl‰ÌÓ„Ó ·ÎÓ͇. èË ˝ÚÓÏ
Ì ÒΉÛÂÚ ÚflÌÛÚ¸ Á‡ ÒÂÚ‚ÓÈ Í‡·Âθ.
òÚÂÍÂ ‚ÒÚ‡‚ËÚ¸ ‚ ̇ÍÓ̘ÌËÍ Ú‡ÍËÏ Ó·‡ÁÓÏ, ˜ÚÓ·˚
˜Û‚ÒÚ‚Ó‚‡ÎÒfl Ë ·˚Î ÒÎ˚¯ÂÌ ˘ÂΘÓÍ.
ÇÓ ‚ÂÏfl ‡·ÓÚ˚ ˜ÂÂÁ ͇·Âθ Ì‚ÓÁÏÓÊÌÓ Á‡flʇڸ
‡ÍÍÛÏÛÎflÚÓ Ò ÔÓÏÓ˘¸˛ Á‡fl‰ÌÓ„Ó ·ÎÓ͇ ÔÓ Ô˘ËÌÂ
ÓÚÒÛÚÒÚ‚Ëfl ˝ÌÂ„ÓÒ̇·ÊÂÌËfl.
èÓÎÌÓ ÓÚÍβ˜ÂÌË ÓÚ ÒÂÚË „‡‡ÌÚËÛÂÚÒfl ÚÓθÍÓ ÔË
‚˚ÌÛÚÓÏ ËÁ ÓÁÂÚÍË ÒÂÚ‚ÓÏ Í‡·ÂÎÂ.
èÂ‰ ͇ʉ˚Ï ËÒÔÓθÁÓ‚‡ÌËÂÏ Á‡„flÁÌÂÌÌ˚ ÔÓ‚Âı-
ÌÓÒÚË ÔË·Ó‡, ‡ Ú‡ÍÊ ҂ÂÚÓ‚Ó‰ Ë Á‡˘ËÚÌ˚È ÍÓÁ˚ÂÍ
ÒΉÛÂÚ ‰ÂÁËÌÙˈËÓ‚‡Ú¸ ËÎË ‡‚ÚÓÍ·‚ËÓ‚‡Ú¸. äÓ-
Ï ÚÓ„Ó, ÒΉÛÂÚ Û‰ÓÒÚÓ‚ÂËÚ¸Òfl,
˜ÚÓ ÏÓ˘ÌÓÒÚ¸ Ò‚ÂÚ‡
ÔË·Ó‡ ‰ÓÒÚ‡ÚӘ̇ ‰Îfl ÔÓÎËÏÂËÁ‡ˆËË Ï‡ÚÂˇ·.
ÑÎfl ˝ÚÓ„Ó ÒΉÛÂÚ „ÛÎflÌÓ ÔÓ‚ÂflÚ¸ Ò‚ÂÚÓ‚Ó‰ ̇
Á‡„flÁÌÂÌËfl Ë ÔÓ‚ÂʉÂÌËfl, ‡ Ú‡ÍÊ ÍÓÌÚÓÎËÓ‚‡Ú¸
ÏÓ˘ÌÓÒÚ¸ ËÁÎÛ˜ÂÌËfl (̇ÔËÏÂ, Ò ÔÓÏÓ˘¸˛ ‡‰ËÓÏÂÚ-
‡ “Bluephase Meter“ ÙËÏ˚ Ivoclar Vivadent).
Ç˚·Ó ÔÓ‡ÏÏ˚ Ë ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË
èÓ„‡ÏÏÛ Ë ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË ÏÓÊÌÓ Ì‡ÒÚ‡Ë-
‚‡Ú¸ Ë̉˂ˉۇθÌÓ. ÑÎfl ‡Á΢Ì˚ı ÔÓ͇Á‡ÌËÈ Î‡ÏÔ‡
Bluephase N ËÏÂÂÚ 3 ÔÓ„‡ÏÏ˚ ÔÓÎËÏÂËÁ‡ˆËË. ë
ÔÓÏÓ˘¸˛ Í·‚Ë¯Ë ‚˚·Ó‡ ÔÓ„‡ÏÏ˚ ÏÓÊÌÓ Á‡‰‡Ú¸
Ê·ÂÏÛ˛ ÔÓ„‡ÏÏÛ ÔÓÎËÏÂËÁ‡ˆËË. àÁÓ·‡ÊÂÌËÂ
̇ ‰ËÒÔΠ·Û‰ÂÚ ËÁÏÂÌflÚ¸Òfl ‚ ÒÓÓÚ‚ÂÚÒÚ‚ËË Ò ‚˚·ÓÓÏ
(ÒÏ. é·ÓÁ̇˜ÂÌËfl ̇ ̇ÍÓ̘ÌËÍÂ). ë Á‡‚Ó‰‡ ÔË·Ó
ÔÓÒÚ‡‚ÎflÂÚÒfl ÒÓ ÒÎÂ‰Û˛˘ËÏË Ì‡ÒÚÓÈ͇ÏË:
HIP (High Power – ‚˚ÒÓ͇fl ÏÓ˘ÌÓÒÚ¸): 10 ÒÂÍÛ̉
LOW (Low Power – ÌËÁ͇fl ÏÓ˘ÌÓÒÚ¸): 10 ÒÂÍÛ̉
SOFí (Soft start – Ïfl„ÍËÈ ÒÚ‡Ú): 15 ÒÂÍÛ̉
Ü·ÂÏÓ ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË Á‡‰‡ÂÚÒfl Í·‚˯ÂÈ
‚˚·Ó‡ ‚ÂÏÂÌË. åÓÊÌÓ ‚˚·‡Ú¸ 5, 10, 15, 20 Ë 30
ÒÂÍÛ̉.
èË ‚˚·Ó ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË ÒΉÛÂÚ
Û˜ËÚ˚‚‡Ú¸ Ú·ӂ‡ÌËfl ËÌÒÚÛ͈ËË Í ËÒÔÓθÁÛÂÏÓÏÛ
χÚÂˇÎÛ. èË ‡·ÓÚÂ Ò ÍÓÏÔÓÁËÚ‡ÏË ÂÍÓÏẨ‡ˆËË
Ó ÔÓÎËÏÂËÁ‡ˆËË ‡ÒÔÓÒÚ‡Ìfl˛ÚÒfl ̇ ‚Ò ˆ‚ÂÚ‡ Ë
ÒÎÓË Ï‡ÚÂˇ· ÚÓ΢ËÌÓÈ Ì ·ÓΠ2 ÏÏ – ÂÒÎË ËÌÓÂ
Ì Û͇Á‡ÌÓ ‚ ËÌÒÚÛ͈ËË. ùÚË ÂÍÓÏẨ‡ˆËË
‰ÂÈÒÚ‚ËÚÂθÌ˚ ‚ ÚÓÏ ÒÎÛ˜‡Â, ÍÓ„‰‡ ÔÓÎËÏÂËÁ‡ˆËfl
ÓÒÛ˘ÂÒÚ‚ÎflÂÚÒfl Ú‡ÍËÏ Ó·‡ÁÓÏ, ˜ÚÓ ‚˚ıÓ‰ÌÓ ÓÍÓ¯ÍÓ
Ò‚ÂÚÓ‚Ó‰‡ ‡ÒÔÓ·„‡ÂÚÒfl ÔflÏÓ ÔÂ‰ ÓÚ‚Âʉ‡ÂÏ˚Ï
χÚÂˇÎÓÏ. ë Û‚Â΢ÂÌËÂÏ ‡ÒÒÚÓflÌËfl ÏÂÊ‰Û ÌËÏË
‰ÓÎÊÌÓ ÒÓÓÚ‚ÂÚÒÚ‚ÂÌÌÓ Û‚Â΢˂‡Ú¸Òfl Ë ‚ÂÏfl
ÔÓÎËÏÂËÁ‡ˆËË. ÖÒÎË ‡ÒÒÚÓflÌË ÏÂÊ‰Û Ò‚ÂÚÓ‚Ó‰ÓÏ Ë
χÚÂˇÎÓÏ ÒÓÒÚ‡‚ÎflÂÚ 9 ÏÏ, ˝ÙÙÂÍÚ˂̇fl ÏÓ˘ÌÓÒÚ¸
ËÁÎÛ˜ÂÌËfl ÒÌËʇÂÚÒfl ÔËÏÂÌÓ Ì‡ 50%, Ú‡Í ˜ÚÓ
ÂÍÓÏẨÛÂÏÓ ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË ‚ ˝ÚÓÏ ÒÎÛ˜‡Â
ÒΉÛÂÚ Û‰‚ÓËÚ¸.
àÒÔÓθÁÓ‚‡ÌË ÔË·Ó‡
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 23
81
èÓ„‡Ïχ ÔÓÎËÏÂËÁ‡ˆËË
1) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl ÚÓ΢ËÌ˚ ÒÎÓfl Ì ·ÓΠ2 ÏÏ, ‡ Ú‡ÍÊ ÂÒÎË ‚
ËÌÒÚÛ͈ËË Í Ï‡ÚÂˇÎÛ ÌÂÚ ‰Û„Ëı Û͇Á‡ÌËÈ (‚ÓÁÏÓÊÌÓ, ̇ÔËÏÂ,
Ò ˆ‚ÂÚ‡ÏË ‰ÂÌÚË̇)
2) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl ÚÓ΢ËÌ˚ ÒÎÓfl Ì ·ÓΠ4 ÏÏ, ‡ Ú‡ÍÊ ÂÒÎË ‚
ËÌÒÚÛ͈ËË Í Ï‡ÚÂˇÎÛ ÌÂÚ ‰Û„Ëı Û͇Á‡ÌËÈ (‚ÓÁÏÓÊÌÓ, ̇ÔËÏÂ,
Ò ˆ‚ÂÚ‡ÏË ‰ÂÌÚË̇)
3) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl ÚÓ΢ËÌ˚ ÒÎÓfl Ì ·ÓΠ3 ÏÏ
4) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl Ò‚ÂÚÓ‚ÓÈ ÔÓÎËÏÂËÁ‡
ˆËË (ËÒÔÓθÁÓ‚‡ÌË ÚÓθÍÓ
Ô‡ÒÚ˚ ·‡Á˚)
5) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl ‰‚ÓÈÌÓÈ ÔÓÎËÏÂËÁ‡ˆËË
èÓ„‡Ïχ HIGH POWER
èÓÒÚÓflÌÌÓ ‚˚ÒÓ͇fl ÏÓ˘ÌÓÒÚ¸
Ò‚ÂÚ‡ ‰Îfl ÔÓÎËÏÂËÁ‡ˆËË
ÔÎÓÏ·ËÓ‚Ó˜Ì˚ı Ë ÙËÍÒËÛ˛˘Ëı
χÚÂˇÎÓ‚ ÔË ‚˚ÔÓÎÌÂÌËË
ÂÒÚ‡‚‡ˆËÈ ÔflÏ˚Ï Ë ÌÂÔflÏ˚Ï
ÏÂÚÓ‰ÓÏ.
èÓ„‡Ïχ SOFT START
ëÚÛÔÂ̘‡ÚÓ ÔÓ‚˚¯ÂÌËÂ
ÏÓ˘ÌÓÒÚË Ò‚ÂÚ‡, ÒÌËÊÂÌË ÒÚÂÒÒ‡
χÚÂˇ· ÔË ÛÒ‡‰ÍÂ,
ÛÏÂ̸¯ÂÌÌÓ ‚˚‰ÂÎÂÌË ÚÂÔ· ‰Îfl
ÔÓÎËÏÂËÁ‡ˆËË ÔÎÓÏ·ËÓ‚Ó˜Ì˚ı
χÚÂˇÎÓ‚.
èÓ„‡Ïχ LOW POWER
èÓÌËÊÂÌ̇fl ÏÓ˘ÌÓÒÚ¸ Ò‚ÂÚ‡ Ò
ÛÏÂ̸¯ÂÌÌ˚Ï ‚˚‰ÂÎÂÌË
ÂÏ ÚÂÔ·
‰Îfl ÔÓÎËÏÂËÁ‡ˆËË ‚ ӷ·ÒÚflı,
·ÎËÁÍËı Í ÔÛθÔÂ, ‡‰„ÂÁË‚Ó‚,
·ÈÌÂÓ‚ Ë ÔÎÓÏ·ËÓ‚Ó˜Ì˚ı
χÚÂˇÎÓ‚ ÔË ÂÒÚ‡‚‡ˆËflı
ÔÓÎÓÒÚÂÈ Í·ÒÒ‡ V.
1.200 mW/cm
2
t [s]
05
650 mW/cm
2
t [s]
0
1.200 mW/cm
2
t [s]
0
650 mW/cm
2
èÎÓÏ·ËÓ‚Ó˜Ì˚ χÚÂˇÎ˚
äÓÏÔÓÁËÚ˚
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
‚Ò Í·ÒÒ˘ÂÒÍË ÍÓÏÔÓÁËÚ˚
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
äÓÏÔÓÏÂ
3)
Compoglass F / Compoglass Flow
çÂÔflÏ˚ ÂÒÚ‡‚‡ˆËË/
îËÍÒËÛ˛˘Ë χÚÂˇÎ˚
Variolink II Base
4)
/ Variolink Veneer / Variolink N Base
4)
/
Variolink N Clear Veneer / Variolink Ultra
5)
/
Dual Cement
5)
/ Variolink II
5)
èӘˠχÚÂˇÎ˚
Helioseal / Helioseal F / Helioseal Clear
Monopaque
MultiCore Flow / Multicore HB
Heliosit Orthodontic
Telio Add-On Flow
Telio Stains
IPS Empress Direct Color
IPS Empress Direct Opaque
ÇÂÏfl ÔÓÎËÏÂËÁ‡ˆËË
10
ÒÂÍ.
15
ÒÂÍ.
10
ÒÂÍ.
20
ÒÂÍ.
ç‡ ÏÏ ÍÂ‡ÏËÍË:
10
ÒÂÍ. ̇ Í‡Ê‰Û˛ ÔÓ‚ÂıÌÓÒÚ¸
10
ÒÂÍ.
20
ÒÂÍ.
20
ÒÂÍ.
10
ÒÂÍ.
15
ÒÂÍ.
10
ÒÂÍ.
10
ÒÂÍ.
20
ÒÂÍ.
ĉ„ÂÁË‚˚
AdheSE / AdheSE One F
ExciTE / ExciTE DSC
Tetric N-Bond / Tetric N-Bond Self-Etch
Heliobond (Syntac)
ÇÂÏÂÌÌ˚ χÚÂˇÎ˚
Telio CS Inlay/Onlay
Systemp.inlay/onlay
Fermit / Fermit N
Telio CS Link / Systemp.link
èӘˠχÚÂˇÎ˚
Heliosit Orthodontic
Vivaglass Liner
ÇÂÏfl ÔÓÎËÏÂËÁ‡ˆËË
10
ÒÂÍ.
10
ÒÂÍ.
10
ÒÂÍ.
10
ÒÂÍ.
20
ÒÂÍ.
̇ Í‡Ê‰Û˛ ÔÓ‚ÂıÌÓÒÚ¸
20
ÒÂÍ.
20
ÒÂÍ.
èÎÓÏ·ËÓ‚Ó˜Ì˚ χÚÂˇÎ˚
äÓÏÔÓÁËÚ˚
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow
‚Ò Í·ÒÒ˘ÂÒÍË ÍÓÏÔÓÁËÚ˚
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric Basic White
äÓÏÔÓÏÂ
3)
Compoglass F / Compoglass Flow
ÇÂÏfl ÔÓÎËÏÂËÁ‡ˆËË
15
ÒÂÍ.
20
ÒÂÍ.
15
ÒÂÍ.
20
ÒÂÍ.
åÓ˘ÌÓÒÚ¸ ËÁÎÛ˜ÂÌËfl
‚ÂÏfl ÓÚ‚ÂʉÂÌËfl ‰Îfl ÍÓÏÔÓÁËÚÓ
‚ÂÏfl ÓÚ‚ÂʉÂÌËfl ‰Îfl Tetric EvoCeram/IPS Empress Direct
1,200 mW/cm
2
15
ÒÂÍ.
10
ÒÂÍ.
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 24
82
îÛÌ͈Ëfl ÒÓı‡ÌÂÌËfl ‰‡ÌÌ˚ı Cure Memory
èÓÒΉÌË ËÒÔÓθÁÓ‚‡ÌÌ˚ ̇ÒÚÓÈÍË ÔÓ„‡ÏÏ˚
ÔÓÎËÏÂËÁ‡ˆËË Ë ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË ‡‚ÚÓχÚË-
˜ÂÒÍË ÒÓı‡Ìfl˛ÚÒfl.
ëÚ‡Ú
äÌÓÔÍÓÈ ëÚ‡Ú ‚Íβ˜‡ÂÚÒfl Ò‚ÂÚÓ‚Ó ËÁÎÛ˜ÂÌËÂ. èÓ
ËÒÚ˜ÂÌËË Á‡‰‡ÌÌÓ„Ó ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË ÔÓ-
„‡Ïχ Á‡‚Â¯‡ÂÚÒfl ‡‚ÚÓχÚ˘ÂÒÍË. èË ÌÂÓ·ıÓ‰ËÏÓ-
ÒÚË Ò‚ÂÚ ÏÓÊÌÓ ÔÂʉ‚ÂÏÂÌÌÓ ÓÚÍβ˜ËÚ¸
ÔÓ‚ÚÓÌ˚Ï Ì‡Ê‡ÚËÂÏ Ì‡ Í·‚Ë¯Û ëÚ‡Ú. é‰ÌÓ‚ÂÏÂÌ-
ÌÓ Ò ‚Íβ˜ÂÌËÂÏ Ò‚ÂÚ‡ ̇˜Ë̇ÂÚ ‡·ÓÚ‡Ú¸ ‚ÂÌÚËÎflÚÓ.
èÓ ÓÍÓ̘‡ÌËË ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË ÓÌ ÔÓ‰ÓÎʇÂÚ
‡·ÓÚ‡Ú¸, Óı·ʉ‡fl ÔË·Ó. èÓ͇ ‚ÂÌÚËÎflÚÓ ‡·ÓÚ‡-
ÂÚ, ‡ÍÍÛÏÛÎflÚÓ ‚˚ÌËχڸ ÌÂθÁfl.
á‚ÛÍÓ‚˚ Ò˄̇Î˚
Ç ÒÎÂ‰Û˛˘Ëı ÒÎÛ˜‡flı Á‚Û˜‡Ú Á‚ÛÍÓ‚˚ Ò˄̇Î˚:
ëÚ‡Ú (ëÚÓÔ)
ä‡Ê‰˚ 10 ÒÂÍÛ̉
ëÏÂ̇ ÔÓ„‡ÏÏ˚ ÔÓÎËÏÂËÁ‡ˆËË
ëÏÂ̇ ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË
ëÓ‰ËÌËÚ¸ ‡ÍÍÛÏÛÎflÚÓ Ò Á‡fl‰Ì˚Ï ·ÎÓÍÓÏ
ìÒÚ‡ÌÓ‚ËÚ¸ ‡ÍÍÛÏÛÎflÚÓ
ëÓÓ·˘ÂÌË ӷ ӯ˷ÍÂ
èË Ê·ÌËË „ÓÏÍÓÒÚ¸ Á‚ÛÍÓ‚˚ı Ò˄̇ÎÓ‚ ÏÓÊÌÓ
„ÛÎËÓ‚‡Ú¸. ÑÎfl ˝ÚÓ„Ó ÔË ‚˚Íβ˜ÂÌÌÓÏ Ò‚ÂÚÓ‚ÓÏ
ËÁÎÛ˜ÂÌËË ÒΉÛÂÚ Ì‡Ê‡Ú¸ ̇ ÒËÌ˛˛ Í·‚Ë¯Û „ÓÏÍÓ-
ÒÚË (ÒÏ. èÂ˜Â̸ ˜‡ÒÚÂÈ) Ë Î‚ÓÈ Í·‚˯ÂÈ ‚˚·Ó‡
‚ÂÏÂÌË ÛÏÂ̸¯ËÚ¸ Á‚ÛÍ ËÎË ÒÓ‚ÒÂÏ ÓÚÍβ˜ËÚ¸ Á‚ÛÍÓ-
‚ÓÈ Ò˄̇Î. ÖÒÎË Ç˚ ıÓÚËÚ ‚Íβ˜ËÚ¸ Á‚ÛÍÓ‚ÓÈ Ò˄̇Î
ËÎË Û‚Â΢ËÚ¸ „ÓÏÍÓÒÚ¸, ÒΉÛÂÚ ÒÌÓ‚‡ ̇ʇڸ ̇
ÒËÌ˛˛ Í·‚Ë¯Û „ÓÏÍÓÒÚË Ë Ô‡‚Û˛ Í·‚Ë¯Û ‚˚·Ó‡
‚ÂÏÂÌË.
åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl
åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl ÒÓı‡ÌflÂÚÒfl
ÔÓÒÚÓflÌÌÓÈ ÔË ‡·ÓÚ ÔË·Ó‡. èË ËÒÔÓθÁÓ‚‡ÌËË
ÔÓÒÚ‡‚ÎflÂÏÓ„Ó Ò ÔË·ÓÓÏ Ò‚ÂÚÓ‚Ó‰‡ 10 ÏÏ ÏÓ˘ÌÓÒÚ¸
ÓÚ͇ÎË·Ó‚‡Ì‡ ̇ 1.200 ÏÇÚ/ÒÏ
2
± 10%
ÖÒÎË ËÒÔÓθÁÛÂÚÒfl ‰Û„ÓÈ Ò‚ÂÚÓ‚Ó‰, ‡ Ì ÚÓÚ, ˜ÚÓ
ÔÓÒÚ‡‚ÎflÂÚÒfl Ò ÔË·ÓÓÏ, ÓÌ ÏÓÊÂÚ Ó͇Á‡Ú¸ ÒÂ¸ÂÁÌÓÂ
‚ÎËflÌË ̇ ÏÓ˘ÌÓÒÚ¸ ËÁÎÛ˜ÂÌËfl ̇ ‚˚ıÓ‰Â.
èË ËÒÔÓθÁÓ‚‡ÌËË Ò‚ÂÚÓ‚Ó‰‡ Ò Ô‡‡ÎÎÂθÌ˚ÏË
ÒÚÂÌ͇ÏË (10 ÏÏ) ‰Ë‡ÏÂÚ˚ ‚ıÓ‰fl˘Â„Ó Ë ‚˚ıÓ‰fl˘Â„Ó
Ò‚ÂÚÓ‚Ó„Ó ÓÍӯ͇ Ó‰Ë̇ÍÓ‚˚. èË ËÒÔÓθÁÓ‚‡ÌËË
ÙÓÍÛÒËÛ˛˘Ëı Ò‚ÂÚÓ‚Ó‰Ó‚ (10>8 ÏÏ, Pin-Point
6>2 ÏÏ) ‰Ë‡ÏÂÚ ̇ ‚ıӉ ·Óθ¯Â, ˜ÂÏ ‰Ë‡ÏÂÚ ̇
‚˚ıÓ‰Â. èË ˝ÚÓÏ Ì‡ ‚˚ıӉ ҂ÂÚ ÙÓÍÛÒËÛÂÚÒfl ̇
Ì·Óθ¯ÓÈ ÔÓ‚ÂıÌÓÒÚË, ˜ÚÓ ÔÓ‚˚¯‡ÂÚ ÏÓ˘ÌÓÒÚ¸
ËÁÎÛ˜ÂÌËfl.
ë‚ÂÚÓ‚Ó‰˚ Pin-Point ıÓÓ¯Ó ÔÓ‰ıÓ‰flÚ ‰Îfl ÚӘ˜ÌÓÈ
ÔÓÎËÏÂËÁ‡ˆËË, ̇ÔËÏÂ, ÙËÍÒ‡ˆËË ‚ËÌËÓ‚ ÔÂ‰
Û‰‡ÎÂÌËÂÏ ËÁÎ˯ÍÓ‚. ÑÎfl Ôӂ‰ÂÌËfl ÔÓÎÌÓ„Ó
ÓÚ‚ÂʉÂÌËfl Ò‚ÂÚÓ‚Ó‰ ÒΉÛÂÚ Á‡ÏÂÌËÚ¸.
àÁÏÂÂÌË ÏÓ˘ÌÓÒÚË Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl
åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl ·ÏÔ˚ Bluephase N Ë
ÔÓÒÚ‡‚ÎflÂÏÓ„Ó Ò ÌÂÈ Ò‚ÂÚÓ‚Ó‰‡ ̇ 10 ÏÏ ÏÓÊÌÓ
ÔÓ‚ÂËÚ¸, ̇ÔËÏÂ, Ò ÔÓÏÓ˘¸˛ Bluephase Meter.
ÖÒÎË ËÁÏÂÂÌÌÓ Á̇˜ÂÌË Ì ÒÓÓÚ‚ÂÚÒÚ‚ÛÂÚ
ÓÊˉ‡ÂÏÓÏÛ:
ÔÓ‚ÂËÚ¸ ‚˚·‡ÌÌÛ˛ ÔÓ„‡ÏÏÛ ÔÓÎËÏÂËÁ‡ˆËË
Ó˜ËÒÚËÚ¸ ÓÚ Á‡„flÁÌÂÌËÈ Ò‚ÂÚÓ‚ÓÈ ÒÂÌÒÓ
‡‰ËÓÏÂÚ‡
ÒÌflÚ¸ Ò‚ÂÚÓ‚Ó‰, ‡ ‚˚ıÓ‰ÌÓ ÓÚ‚ÂÒÚË ҂ÂÚ‡ ̇
̇ÍÓ̘ÌËÍ ÔÓ˜ËÒÚËÚ¸ ‚‡ÚÌÓÈ Ô‡ÎÓ˜ÍÓÈ,
ÒÏÓ˜ÂÌÌÓÈ ÒÔËÚÓÏ.
‚ÓÁÏÓÊÌÓ, Á‡„flÁÌËÎÒfl Ò‚ÂÚÓ‚Ó‰ Ë Â„Ó ÒΉÛÂÚ
ÔÓ˜ËÒÚËÚ¸ (ÒÏ. ‡Á‰ÂÎ íÂıÓ·ÒÎÛÊË‚‡ÌËÂ Ë Ó˜ËÒÚ͇)
ÂÒÎË Ò‚ÂÚÓ‚Ó‰ ÔÓ‚ÂʉÂÌ, ÌÂÓ·ıÓ‰ËÏÓ Á‡ÏÂÌËÚ¸ „Ó
̇ ÌÓ‚˚È
ÖÒÎË ˝ÚË ÏÂ˚ Ì ÔËÌÂÒÎË Ê·ÂÏÓ„Ó ÂÁÛθڇڇ,
Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û.
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 25
83
ÑÎfl ÎÛ˜¯Â„Ó Òӷβ‰ÂÌËfl Ú·ӂ‡ÌËÈ „Ë„ËÂÌ˚
ÂÍÓÏẨÛÂÚÒfl ‰Îfl Í‡Ê‰Ó„Ó Ô‡ˆËÂÌÚ‡ ËÒÔÓθÁÓ‚‡Ú¸
Ó‰ÌÓ‡ÁÓ‚˚ ˜ÂıÎ˚. ᇄflÁÌÂÌÌ˚ ÔÓ‚ÂıÌÓÒÚË ÔË·Ó‡,
‡ Ú‡ÍÊ Á‡˘ËÚÌ˚È ÍÓÁ˚ÂÍ ÒΉÛÂÚÔÂ‰ ͇ʉ˚Ï
ËÒÔÓθÁÓ‚‡ÌËÂÏ ‰ÂÁËÌÙˈËÓ‚‡Ú¸ (̇ÔËÏÂ, FD333,
FD366/Dürr Dental, Incidin Liquid/Ecolab), ‚ ÒÎÛ˜‡Â ÂÒÎË ÌÂ
·˚ÎË ËÒÔÓθÁÓ‚‡Ì˚ Á‡˘ËÚÌ˚ ˜ÂıÎ˚ Ò‚ÂÚÓ‚Ó‰ ‰ÓÎÊÂÌ
·˚Ú¸ ÒÚÂÂÎËÁÓ‚‡Ì ÔÂ‰ ͇ʉ˚Ï ÔËÏÂÌÂÌËÂÏ. èË
‡·ÓÚ‡ı ÔÓ Ó˜Ë
ÒÚÍ ÔË·Ó‡ ̇ ̇ÍÓ̘ÌËÍ, Á‡fl‰Ì˚È
·ÎÓÍ, ‡ ‚ ÓÒÓ·ÂÌÌÓÒÚË Ì‡ ÒÂÚ‚ÓÈ ·ÎÓÍ Ì ‰ÓÎÊÌ˚
ÔÓÔ‡‰‡Ú¸ ÌË͇ÍË ÊˉÍÓÒÚË ËÎË ‰Û„Ë χÚÂˇÎ˚
(ÓÔ‡ÒÌÓÒÚ¸ ÔÓ‡ÊÂÌËfl ˝ÎÂÍÚ˘ÂÒÍËÏ ÚÓÍÓÏ). èË Ó˜ËÒÚÍÂ
Á‡fl‰Ì˚È ·ÎÓÍ ÌÂÓ·ıÓ‰ËÏÓ ÓÚÍβ˜ËÚ¸ ÓÚ ˝ÎÂÍÚÓÒÂÚË.
äÓÔÛÒ
ç‡ÍÓ̘ÌËÍ Ë ‰ÂʇÚÂθ ̇ÍÓ̘ÌË͇ ÔÓÚË‡Ú¸ Ó·˚˜Ì˚Ï
‰ÂÁËÌÙÂ͈ËÓÌÌ˚Ï Ò‰ÒÚ‚ÓÏ, Ì ÒÓ‰Âʇ˘ËÏ ‡Î¸‰Â„ˉӂ.
ç ËÒÔÓθÁÓ‚‡Ú¸ ÌË͇ÍËı ‡„ÂÒÒË‚Ì˚ı ‰ÂÁËÌÙÂ͈ËÓÌÌ˚ı
Ò‰
ÒÚ‚ (̇ÔËÏÂ, ‡ÒÚ‚Ó˚ ̇ ÓÒÌÓ‚Â ‡ÔÂθÒËÌÓ‚Ó„Ó
χÒ· ËÎË ‡ÒÚ‚Ó˚ Ò ÒÓ‰ÂʇÌËÂÏ ˝Ú‡ÌÓ· Ò‚˚¯Â 40%),
‡ÒÚ‚ÓËÚÂÎÂÈ (̇ÔËÏÂ, ‡ˆÂÚÓ̇) ËÎË ÓÒÚ˚ı Ô‰ÏÂÚÓ‚,
ÍÓÚÓ˚ ÏÓ„ÛÚ ÔÓ‚‰ËÚ¸ ËÎË ÔÓˆ‡‡Ô‡Ú¸ Ô·ÒÚχÒÒÛ.
ᇄflÁÌÂÌÌ˚ Ô·ÒÚχÒÒÓ‚˚ ˜‡ÒÚË Ó˜Ë˘‡Ú¸ Ï˚θÌ˚Ï
‡ÒÚ‚ÓÓÏ.
ë‚ÂÚÓ‚Ó‰
ÔÂ‰ Ó˜ËÒÚÍÓÈ Ë/ËÎË ‰ÂÁËÌÙÂ͈ËÂÈ Ò‚ÂÚÓ‚Ó‰‡ Â„Ó ÒΉÛÂÚ
Ô‰‚‡ËÚÂθÌÓ Ó·‡-·ÓÚ‡Ú¸. ùÚÓ Ô‡‚ËÎÓ ‰ÂÈÒÚ‚ÛÂÚ Ë
ÒÎÛ˜‡Â χ¯ËÌÌÓÈ, Ë ‚
ÒÎÛ˜‡Â Û˜ÌÓÈ Ó˜ËÒÚÍË Ë ‰ÂÁËÌÙÂ͈ËË.
è‰‚‡ËÚÂθ̇fl Ó·‡·ÓÚ͇
ìÒÚ‡ÌËÚ „Û·˚ Á‡„flÁÌÂÌËfl Ò‡ÁÛ ÔÓÒΠÔËÏÂÌÂÌËfl,
Ò‡ÏÓ ÔÓÁ‰Ì ˜ÂÂÁ ‰‚‡ ˜‡Ò‡ ÔÓÒΠÔËÂχ Ô‡ˆËÂÌÚ‡.
í˘‡ÚÂθÌÓ ÔÓÏÓÈÚ ҂ÂÚÓ‚Ó‰ ÔÓ‰ ÔÓÚÓ˜ÌÓÈ ‚Ó‰ÓÈ (ÌÂ
ÏÂÌ 10 ÒÂÍÛ̉). åÓÊÌÓ Ú‡ÍÊ ËÒÔÓθÁÓ‚‡Ú¸ ÔÓ‰ıÓ‰fl˘ÂÂ
‰ÂÁËÌÙÂ͈ËÓÌÌÓ Ò‰ÒÚ‚Ó, Ì ÒÓ‰Âʇ˘Â ‡Î¸‰Â„ˉӂ,
˜ÚÓ·˚ ÔÂÔflÚÒÚ‚Ó‚‡Ú¸ ÔËÒ˚ı‡Ì˲ ÍÓ‚Ë.
óÚÓ·˚ Û‰‡ÎËÚ¸ Á‡„flÁÌÂÌËfl ‚Û˜ÌÛ˛, ËÒÔÓ
θÁÛÈÚ ÎÛ˜¯Â
‚ÒÂ„Ó Ïfl„ÍÛ˛ ˘ÂÚÍÛ ËÎË Ïfl„ÍÛ˛ Ò‡ÎÙÂÚÍÛ. èÓÎËÏÂËÁÓ-
‚‡ÌÌ˚È ÍÓÏÔÓÁËÚ ÏÓÊÌÓ Û‰‡ÎËÚ¸ ÒÔËÚÓÏ ËÎË Ô·ÒÚ-
χÒÒÓ‚˚Ï ¯Ô‡ÚÂÎÂÏ. ç ËÒÔÓθÁÓ‚‡Ú¸ ÌË͇ÍËı ÓÒÚ˚ı
Ô‰ÏÂÚÓ‚, ÍÓÚÓ˚ ÏÓ„ÛÚ ÔÓˆ‡‡Ô‡Ú¸ ÔÓ‚ÂıÌÓÒÚ¸.
é˜ËÒÚ͇ Ë ‰ÂÁËÌÙÂ͈Ëfl
ÑÎfl Ó˜ËÒÚÍË ÔÓÎÓÊËÚ ҂ÂÚÓ‚Ó‰ ‚ Ó˜Ë˘‡˛˘ËÈ ‡ÒÚ‚Ó,
Ú‡Í, ˜ÚÓ·˚ ÓÌ ÔÓÎÌÓÒÚ¸˛ ·˚Î ÔÓÍ˚Ú ‡ÒÚ‚ÓÓÏ (ÛθÚ‡-
Á‚ÛÍÓ‚‡fl ‚‡Ì̇ ËÎË ÓÒÚÓÓÊÌÓÂ Ó˜Ë˘ÂÌË Ïfl„ÍÓÈ ˘ÂÚÍÓÈ
ÏÓ„ÛÚ ÛÒËÎËÚ¸
‰ÂÈÒÚ‚ËÂ). êÂÍÓÏẨÛÂÚÒfl ÌÂÈÚ‡Î¸ÌÓÂ
ÙÂÏÂÌÚ‡ˆËÓÌÌÓ ÏÓ˛˘Â Ò‰ÒÚ‚Ó. é·‡˘‡ÈÚÂ
‚ÌËχÌËÂ, ˜ÚÓ Ò‰ÒÚ‚Ó, ËÒÔÓθÁÛÂÏÓ ‰Îfl Ó˜ËÒÚÍË Ë
‰ÂÁËÌÙÂÍ-ˆËË Ì ÒÓ‰ÂÊËÚ
– Ó„‡Ì˘ÂÒÍËÂ, ÏËÌÂ‡Î¸Ì˚Â Ë ÓÍËÒÎfl˛˘Ë ÍËÒÎÓÚ˚
(ÏËÌËχθÌÓ ‰ÓÔÛÒÚËÏÓ Á̇˜Â-ÌË pH 5,5),
– ˘ÂÎÓ˜Ë (χÍÒËχθÌÓ ‰ÓÔÛÒÚËÏÓ Á̇˜ÂÌË pH 8,5),
– ÓÍËÒÎfl˛˘Ë Ò‰ÒÚ‚‡ (̇ÔËÏÂ, ÔÂÓÍÒˉ ‚Ó‰ÓÓ‰‡)
èÓÒΠ˝ÚÓ„Ó ‚˚̸Ú ҂ÂÚÓ‚Ó‰ ËÁ ‡ÒÚ‚Ó‡ Ë Ú˘‡ÚÂθÌÓ
ÔÓ
ÏÓÈÚ ÔÓ‰ ÔÓÚÓ˜ÌÓÈ ‚Ó‰ÓÈ (Ì ÏÂÌ 10 ÒÂÍÛ̉).
ùÙÙÂÍÚË‚ÌÓÈ ‡Î¸ÚÂ̇ÚË‚ÓÈ fl‚ÎflÂÚÒfl Ó˜ËÒÚ͇
ÚÂÏÓ‰ÂÁËÌ-ÙÂÍÚÓÂ.
ëÚÂËÎËÁ‡ˆËfl
àÌÚÂÌÒ˂̇fl Ó˜ËÒÚ͇ Ë ÒÚÂËÎËÁ‡ˆËfl ÌÂÓ·ıÓ‰ËÏ˚ ‰Îfl
ÚÓ„Ó, ˜ÚÓ·˚ ÔÓÒÎÂ‰Û˛˘‡fl ÒÚÂ-ËÎËÁ‡ˆËfl ·˚· ˝ÙÙÂÍÚË‚ÌÓÈ.
èÓʇÎÛÈÒÚ‡, ÔÓθÁÛÈÚÂÒ¸ ËÒÍβ˜ËÚÂθÌÓ ÚÓθÍÓ Ô‡Ó-‚ÓÈ
ÒÚÂËÎËÁ‡ˆËÂÈ. ÇÂÏfl ÒÚÂËÎËÁ‡ˆËË (‚ÂÏfl ˝ÍÒÔÓÁˈËË
ÔË ÚÂÏÔÂ‡ÚÛ ÒÚÂËÎË-Á‡ˆËË) ÒÓÒÚ‡‚ÎflÂÚ 4 ÏËÌÛÚ˚ ÔË
134°C; ‰‡‚ÎÂÌË
 ‰ÓÎÊÌÓ ÒÓÒÚ‡‚ÎflÚ¸ 2 ·‡‡. èÓÒÛ¯Ë-ÚÂ
ÒÚÂËÎËÁÓ‚‡ÌÌ˚È Ò‚ÂÚÓ‚Ó‰ ËÎË Ò ÔÓÏÓ˘¸˛ ÒÔˆˇθÌÓÈ
ÔÓ„‡ÏÏ˚ ÔÓÒۯ˂‡ÌËfl ‚‡¯Â„Ó Ô‡Ó‚Ó„Ó ‡‚ÚÓÍ·‚‡ ËÎË
Ò ÔÓÏÓ˘¸˛ „Ófl˜Â„Ó ‚ÓÁ‰Ûı‡. ë‚ÂÚÓ‚Ó‰ ÔÓ ÂÁÛθڇڇÏ
ÚÂÒÚÓ‚ ‚˚‰ÂÊË‚‡ÂÚ ‰Ó 200 ˆËÍÎÓ‚ ÒÚÂËÎËÁ‡ˆËË.
á‡ÚÂÏ ÔÓ‚Â¸Ú ҂ÂÚÓ‚Ó‰ ̇ ̇΢ˠÔÓ‚ÂʉÂÌËÈ. ÑÎfl
˝ÚÓ„Ó ÌÂÓ·ıÓ‰ËÏÓ ÔÓÒÏÓÚÂÚ¸ Â„Ó Ì‡ ÔÓÒ‚ÂÚ. ÖÒÎË ÔË
˝ÚÓÏ ‚ˉÌ˚ ÓÚ‰ÂθÌ˚ ˜ÂÌ˚ Ò„ÏÂÌÚ˚ – Á̇˜ËÚ, ˝ÚÓ
ÏÂÒÚ‡ ÔÂ
ÂÎÓÏÓ‚ ÒÚÂÍÎÓ‚ÓÎÓÍ̇, Ò‚ÂÚÓ‚Ó‰ ÒΉÛÂÚ
Á‡ÏÂÌËÚ¸ ̇ ÌÓ‚˚È.
äÓÌÚ‡ÍÚ˚ ‡ÍÍÛÏÛÎflÚÓ‡
ÑÎfl Ó·ÂÒÔ˜ÂÌËfl ÔÓÒÚÓflÌÌÓÈ ıÓÓ¯ÂÈ ÔÓ‚Ó‰ËÏÓÒÚË
ÍÓÌÚ‡ÍÚ˚ ‡ÍÍÛÏÛÎflÚÓ‡ ÒΉÛÂÚ Á‡˘Ë˘‡Ú¸ ÓÚ Á‡„flÁÌÂÌËÈ
(̇ÔËÏÂ,ÓÒÚ‡ÚÍÓ‚ ÍÓÏÔÓÁËÚ‡). ÑÎfl ˝ÚÓ„Ó ÍÓÌÚ‡ÍÚ˚
ÒΉÛÂÚ „ÛÎflÌÓ Ó˜Ë˘‡Ú¸ ‚ ‡Ï͇ı Ó·˚˜ÌÓÈ ‰ÂÁËÌÙÂ͈ËË
(ÔÓÒÎÂ Í‡Ê‰Ó„Ó Ô‡ˆËÂÌÚ‡).
ìÚËÎËÁ‡ˆËfl
èË·Ó ÌÂθÁfl ÛÚËÎËÁËÓ‚‡Ú¸ Ò Ó·˚˜Ì˚Ï
·˚ÚÓ‚˚Ï ÏÛÒÓÓÏ. ç„ӉÌ˚ ÔË·Ó˚ ‰Îfl
ÔÓÎËÏÂ
ËÁ‡ˆËË Ë ‡ÍÍÛÏÛÎflÚÓ˚ ÒΉÛÂÚ
ÛÚËÎËÁËÓ‚‡Ú¸ ‚ ÒÓÓÚ‚ÂÚÒÚ‚ËË Ò
̇ˆËÓ̇θÌ˚ÏË Á‡ÍÓÌÓ‰‡ÚÂθÌ˚ÏË ÌÓχÏË.
ÄÍÍÛÏÛÎflÚÓ˚ ÌË ‚ ÍÓÂÏ ÒÎÛ˜‡Â Ì ·Ó-Ò‡Ú¸ ‚
Ó„Ó̸!
íÂıÓ·ÒÎÛÊË‚‡ÌËÂ Ë Ó˜ËÒÚ͇
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 26
84
óÚÓ ‰Â·ڸ, ÂÒÎË …?
ëËÏ‚ÓÎ è˘Ë̇ ìÒÚ‡ÌÂÌËÂ
ÑÂÙÂÍÚ ˝ÎÂÍÚÓÌËÍË ‚ ̇ÍÓ̘ÌË-
Í ËÎË ‡ÍÍÛÏÛÎflÚÓÂ
èË·Ó ÔÂ„ÂÎÒfl ËÎË
ÔÂÂÓı·‰ËÎÒfl
ÑÂÙÂÍÚ ˝ÎÂÍÚÓÌËÍË
‡ÍÍÛÏÛÎflÚÓ‡
ÄÍÍÛÏÛÎflÚÓ ‡ÁflÊÂÌ
á‡flʇ˛˘Ë ÍÓÌÚ‡ÍÚ˚ Á‡„flÁÌÂÌ˚
ÄÍÍÛÏÛÎflÚÓ ‚ ·ÂÁÓÔ‡ÒÌÓÏ ÂÊË-
Ï (‚ÒÚÓÂÌÌ˚È Á‡˘ËÚÌ˚È
‚˚Íβ˜‡ÚÂθ)
ëÂÚ‚ÓÈ ·ÎÓÍ Ì ÔÓ‰Íβ˜ÂÌ ËÎË
‰ÂÙÂÍÚ ÒÂÚÂ‚Ó„Ó ·ÎÓ͇.
ÄÍÍÛÏÛÎflÚÓ ‚˚ÌÛÚ¸ Ë ÒÌÓ‚‡ ‚ÒÚ‡‚ËÚ¸. ÖÒÎË Ó¯Ë·Í‡
ÒÓı‡ÌflÂÚÒfl, Á‡ÏÂÌËÚ¸ ‡ÍÍÛÏÛÎflÚÓ ÒÂÚ‚˚Ï ÔËÚ‡ÌËÂÏ
(ÙÛÌ͈Ëfl Click & Cure).
ÖÒÎË Ó¯Ë·Í‡ ÒÌÓ‚‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË
‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û.
чڸ ÔË·ÓÛ Óı·‰ËÚ¸Òfl (ËÎË ÔË ÔÂÂÓı·ʉÂÌÌÓÏ
ÔË·Ó ‰Óʉ‡Ú¸Òfl, ÔÓ͇ ÓÌ ‰ÓÒÚË„ÌÂÚ ÍÓÏ̇ÚÌÓÈ
ÚÂÏÔÂ‡ÚÛ˚) Ë ˜ÂÂÁ ÌÂÍÓÚÓÓ ‚ÂÏfl ÔÓÔ˚Ú‡Ú¸Òfl ¢ ‡Á.
ÖÒÎË ÌÂÔÓ·‰Í‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚
ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û.
ÄÍÍÛÏÛÎflÚÓ ‚˚ÌÛÚ¸ Ë ÒÌÓ‚‡ ‚ÒÚ‡‚ËÚ¸. ÖÒÎË Ó¯Ë·Í‡
ÒÓı‡ÌflÂÚÒfl, ‚ÒÚ‡‚ËÚ¸ ̇ÍÓ̘ÌËÍ ‚ Á‡fl‰Ì˚È ·ÎÓÍ. ÖÒÎË
ӯ˷͇ ÒÓı‡ÌflÂÚÒfl, Á‡ÏÂÌËÚ¸ ‡ÍÍÛÏÛÎflÚÓ ÒÂÚ‚˚Ï
ÔËÚ‡ÌËÂÏ (ÙÛÌ͈Ëfl Click & Cure).
é·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û.
ç‡ÍÓ̘ÌËÍ ‚ÒÚ‡‚ËÚ¸ ‚ Á‡fl‰Ì˚È ·ÎÓÍ. ç‡ ‰ËÒÔΠÔÓfl‚ËÚÒfl
Í‡ÚÍÓ‚ÂÏÂÌ̇fl ̇‰ÔËÒ¸ „Charging“ ÔÂʉ ˜ÂÏ ‰ËÒÔÎÂÈ
ÔÂÂȉÂÚ ‚ ÂÊËÏ Stand-by. ÖÒÎË ‡ÍÍÛÏÛÎflÚÓ ÌÂ
Á‡flʇÂÚÒfl, ÒΉÛÂÚ ÔÓ˜ËÒÚËÚ¸ ÍÓÌÚ‡ÍÚ˚.
ÖÒÎË ÌÂÔÓ·‰Í‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚
ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û.
èÓÍÓÌÚÓÎËÓ‚‡Ú¸, Ô‡‚ËθÌÓ ÎË ‚ÒÚ‡‚ÎÂÌ Ì‡ÍÓ̘ÌËÍ ‚
Á‡fl‰Ì˚È ·ÎÓÍ. èÓ˜ËÒÚËÚ¸ ÍÓÌÚ‡ÍÚ˚ ‡ÍÍÛÏÛÎflÚÓ‡.
ÑÎfl ËÌÙÓχˆËË: ç‡ ‰ËÒÔΠÔÓfl‚ËÚÒfl Í‡ÚÍÓ‚ÂÏÂÌ̇fl
̇‰ÔËÒ¸ “Charging“ ÔÂʉ ˜ÂÏ ‰ËÒÔÎÂÈ ÔÂÂȉÂÚ ‚ ÂÊËÏ
Stand-by.
ÖÒÎË Ó˜ËÒÚ͇ ÍÓÌÚ‡ÍÚÓ‚ ÛÒÔÂı‡ Ì ÔËÌÂÒ·, ‡ÍÍÛÏÛÎflÚÓ
ÒΉÛÂÚ ‚˚ÌÛÚ¸. á‡ÚÂÏ ‡ÍÍÛÏÛÎflÚÓ ÓÚ‰ÂθÌÓ ÓÚ
̇ÍÓ̘ÌË͇ Á‡flʇڸ ‚ Á‡fl‰ÌÓÏ ·ÎÓÍ ÔËÏÂÌÓ
10 ÏËÌÛÚ. ÖÒÎË ÌÂÔÓ·‰Í‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í
ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û.
èÓÍÓÌÚÓÎËÓ‚‡Ú¸, Ô‡‚ËθÌÓ ÎË ÔÓ‰ÒÓ‰ËÌÂÌ ÒÂÚ‚ÓÈ
·ÎÓÍ Í Á‡fl‰ÌÓÏÛ ·ÎÓÍÛ, ËÎË ÔÓ‰ÒÓ‰ËÌÂÌ ÎË ÒÂÚ‚ÓÈ ·ÎÓÍ
Í ÒÂÚË ÒÂÚ‚˚Ï Í‡·ÂÎÂÏ (ÔË Ô‡‚ËθÌÓÈ ‡·ÓÚ ̇
ÒÂÚ‚ÓÏ ·ÎÓÍ „ÓËÚ ÁÂÎÂ̇fl ·ÏÔӘ͇).
ÖÒÎË ÌÂÔÓ·‰Í‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚
ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û.
èË Á‡fl‰-
Í ̇ ‰ËÒ-
ÔÎÂÂ ÌÂ
ÔÓfl‚ÎflÂÚÒfl
ÌË͇ÍËı
ËÁÓ·‡ÊÂ-
ÌËÈ
ë‚ÂÚÓ‰ËÓ‰
Á‡fl‰ÌÓ„Ó
·ÎÓ͇ ÌÂ
„ÓËÚ
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 27
85
ɇ‡ÌÚËfl /
ÑÂÈÒÚ‚Ëfl ‚ ÒÎÛ˜‡Â ÂÏÓÌÚ‡
ëÔˆËÙË͇ˆËfl
ɇ‡ÌÚËÈÌ˚È ÒÓÍ ‰Îfl ÔË·Ó‡ Bluephase N ÒÓÒÚ‡‚ÎflÂÚ 3
„Ó‰‡ ÒÓ ‰Ìfl ÔÓÍÛÔÍË (̇ ‡ÍÍÛÏÛÎflÚÓ – 1 „Ó‰).
Ç ÒÎÛ˜‡Â ÌÂÔÓ·‰ÓÍ, Ó·ÛÒÎÓ‚ÎÂÌÌ˚ı ‰ÂÙÂÍÚÓÏ
χÚÂˇ· ËÎË Ó¯Ë·ÍÓÈ ÔË ËÁ„ÓÚÓ‚ÎÂÌËË, „‡‡ÌÚËfl
Ó·ÂÒÔ˜˂‡ÂÚ ·ÂÒÔ·ÚÌ˚È ÂÏÓÌÚ ‡ÔÔ‡‡Ú‡. ë‚Âı
˝ÚÓ„Ó „‡‡ÌÚËfl Ì ‰‡ÂÚ Ô‡‚‡ ̇ ‚ÓÁÏ¢ÂÌËÂ
χÚÂˇθÌÓ„Ó ËÎË ÏÓ‡Î¸ÌÓ„Ó Û˘Â·‡. èË ˝ÚÓÏ
ÔË·Ó ‰ÓÎÊÂÌ ËÒÔÓθÁÓ‚‡Ú¸Òfl ËÒÍβ˜ËÚÂθÌÓ ÚÓθÍÓ
ÔÓ Ì‡Á̇˜ÂÌ˲. ã˛·Ó ËÒÔÓθÁÓ‚‡ÌË ‚ ËÌ˚ı ˆÂÎflı
fl‚ÎflÂÚÒfl ËÒÔÓθÁÓ‚‡ÌËÂÏ Ì ÔÓ Ì‡Á̇˜ÂÌ˲ – Á‡
ÔÓÎÛ˜ÂÌÌ˚ ÂÁÛθڇÚ˚ ÔÓËÁ‚Ó‰ËÚÂθ Ì ÌÂÒÂÚ
ÓÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚË Ë Ì ӷÂÒÔ˜˂‡ÂÚ „‡‡ÌÚ˲. ä Ú‡ÍËÏ
ÒÎÛ˜‡flÏ ÓÚÌÓÒflÚÒfl:
Û˘Â·, ̇ÌÂÒÂÌÌ˚È ‚ ÂÁÛθڇÚ ÌÂÍÓÂÍÚÌÓ„Ó
Ó·‡˘ÂÌËfl Ò ÔË·ÓÓÏ. éÒÓ·ÂÌÌÓ ˝ÚÓ ÓÚÌÓÒËÚÒfl Í
ÌÂÔ‡‚ËθÌÓÏÛ ı‡ÌÂÌ˲ ‡ÍÍÛÏÛÎflÚÓÓ‚ (ÒÏ.
íÂıÌ˘ÂÒÍË ‰‡ÌÌ˚Â: ìÒÎÓ‚Ëfl Ú‡ÌÒÔÓÚËÓ‚ÍË Ë
ı‡ÌÂÌËfl).
ÔÓ‚ÂʉÂÌË ‰ÂÚ‡ÎÂÈ, ÍÓÚÓ˚ ÔÓ‰ÎÂÊ‡Ú ËÁÌÓÒÛ ‚
ÂÁÛθڇÚ ÌÓχθÌÓÈ ‡·ÓÚ˚ (̇ÔËÏÂ,
‡ÍÍÛÏÛÎflÚÓ).
ÔÓ‚ÂʉÂÌËfl ‚ ÂÁÛθڇÚ ‚̯ÌËı ‚ÓÁ‰ÂÈÒÚ‚ËÈ,
̇ÔËÏÂ, Û‰‡‡, Ô‡‰ÂÌËfl ̇ ÔÓÎ
ÔÓ‚ÂʉÂÌËfl ËÁ-Á‡ ÌÂÍÓÂÍÚÌÓÈ ÛÒÚ‡ÌÓ‚ÍË ÎË·Ó
ËÌÒÚ‡ÎÎflˆËË ÔË·Ó‡
ÔÓ‚ÂʉÂÌËfl, ÔÓÎÛ˜ÂÌÌ˚ ‚ ÂÁÛθڇÚ ÔÓ‰Íβ˜Â-
ÌËfl ÔË·Ó‡ Í ÒÂÚË Ò Ì‡ÔflÊÂÌËÂÏ Ë ˜‡ÒÚÓÚÓÈ,
ÓÚ΢Ì˚ÏË ÓÚ Û͇Á‡ÌÌ˚ı ̇ Ú‡·Î˘Í ÔË·Ó‡.
ÔÓ‚ÂʉÂÌËfl, ÔÓÎÛ˜ÂÌÌ˚ ‚ ÂÁÛθڇÚ ÂÏÓÌÚÌ˚ı
‡·ÓÚ ÎË·Ó ËÁÏÂÌÂÌËÈ ‚ ÔË·ÓÂ, ÍÓÚÓ˚ ·˚ÎË
ÔÓËÁ‚‰ÂÌ˚ Ó„‡ÌËÁ‡ˆËflÏË, Ì Ëϲ˘ËÏË
ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘ÂÈ ‡‚ÚÓËÁ‡ˆËË.
ÖÒÎË ÒÎÛ˜‡È ÔËÁÌ‡Ì „‡‡ÌÚËÈÌ˚Ï, ‚ÂÒ¸ ÔË·Ó
Ò·Ó (̇ÍÓ̘ÌËÍ, Á‡fl‰Ì˚È ·ÎÓÍ, ‡ÍÍÛÏÛÎflÚÓ,
ÒÂÚ‚ÓÈ Í‡·Âθ, ÒÂÚ‚ÓÈ ·ÎÓÍ) ÒΉÛÂÚ ÔÓÒ·ڸ ‚ÏÂÒÚÂ
Ò ‰ÓÍÛÏÂÌÚÓÏ Ó· ÓÔ·Ú ÔË·Ó‡ ‚ ÓË„Ë̇θÌÓÈ
ÛÔ‡ÍÓ‚ÍÂ Ò ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘ËÏË Í‡ÚÓÌÌ˚ÏË ‚Í·‰Í‡ÏË
(‰ÓÒÚ‡‚ÍÛ ÓÔ·˜Ë‚‡ÂÚ ÔÓÚ·ËÚÂθ) ÔÓ‰‡‚ˆÛ ËÎË
̇ÔflÏÛ˛ ̇ Vivadent Ivoclar.
ã˛·˚ ÂÏÓÌÚÌ˚ ‡·ÓÚ˚ ÏÓ„ÛÚ ÔÓËÁ‚Ó‰ËÚ¸Òfl ÚÓθÍÓ
Í‚‡ÎËÙˈËÓ‚‡ÌÌ˚Ï ÔÂÒÓ̇ÎÓÏ ÒÂ‚ËÒÌÓÈ ÒÎÛÊ·˚,
Ëϲ˘ÂÈ ‡‚ÚÓËÁ‡ˆË˛ ÓÚ Ivoclar Vivadent. Ç ÒÎÛ˜‡Â
‚ÓÁÌËÍÌÓ‚ÂÌËfl ‰ÂÙÂÍÚ‡, ÍÓÚÓ˚È Ì ÏÓÊÂÚ ·˚Ú¸
ÛÒÚ‡ÌÂÌ Ç‡ÏË, Ó·‡ÚËÚÂÒ¸, ÔÓʇÎÛÈÒÚ‡, Í Ç‡¯ÂÏÛ
ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û (‡‰ÂÒ‡ Ç˚ ̇ȉÂÚÂ
̇ Ó·ÎÓÊÍ ËÌÒÚÛ͈ËË). óÂÚÍÓ ÓÔËÒ‡ÌË ‰ÂÙÂÍÚ‡
ËÎË Ó·ÒÚÓflÚÂθÒÚ‚, ÍÓÚÓ˚ ÔË‚ÂÎË Í ‰ÂÙÂÍÚÛ,
ӷ΄˜‡˛Ú ÔÓËÒÍ ÌÂÔÓ·‰ÍË. èÓʇÎÛÈÒÚ‡, ÔËÎÓÊËÚÂ
˝ÚÓ ÓÔËÒ‡ÌËÂ Í ‡ÔÔ‡‡ÚÛ.
ëÔˆËÙË͇ˆËfl
1 Á‡fl‰Ì˚È ·ÎÓÍ, 1 ÒÂÚ‚ÓÈ Í‡·Âθ, 1 ÒÂÚ‚ÓÈ ·ÎÓÍ, 1 ̇ÍÓ̘ÌËÍ,
1 ‡ÍÍÛÏÛÎflÚÓ (Li-Po), 1 Ò‚ÂÚÓ‚Ó‰ 10 ÏÏ, 3 Á‡˘ËÚÌ˚ı ˝Í‡Ì‡,
1 ËÌÒÚÛ͈Ëfl
èË̇‰ÎÂÊÌÓÒÚË
ÑÎfl LJ¯ÂÈ Î‡ÏÔ˚ Bluephase N ‚˚ÔÛÒ͇˛ÚÒfl ÒÎÂ‰Û˛˘Ë ÔË̇‰ÎÂÊÌÓÒÚË:
ÄÚ.‹ ç‡ËÏÂÌÓ‚‡ÌËÂ
656196 ë‚ÂÚÓ‚Ó‰ 10 ÏÏ ˜ÂÌ˚È (N)
60853
8 ë‚ÂÚÓ‚Ó‰ Pin-point 6>2 ÏÏ ˜ÂÌ˚È
551756 ᇢËÚÌ˚È ÍÓÁ˚ÂÍ (ÔÂ‚ÂÌÛÚ˚È ÍÓÌÛÒ)
592496 ᇢËÚÌ˚È ˝Í‡Ì
659956 ÄÍÍÛÏÛÎflÚÓ Bluephase N
659681 ç‡ÍÓ̘ÌËÍ Bluephase N
607922 Bluephase Meter
608554 Bluephase Á‡˘ËÚÌ˚ ˜ÂıÎ˚
íÂıÌ˘ÂÒÍË ‰‡ÌÌ˚Â
ꇷӘ ̇ÔflÊÂÌË 5 VDC
Á‡fl‰ÌÓ„Ó ·ÎÓ͇
ꇷӘ ̇ÔflÊÂÌË 3.7 VDC Ò ‡ÍÍÛÏÛÎflÚÓÓÏ
̇ÍÓ̘ÌË͇ 5 VDC Ò ÒÂÚ‚˚Ï ·ÎÓÍÓÏ
ëÂÚ‚ÓÈ ·ÎÓÍ 100–240 VAC / 50-60 Ɉ /
χÍÒ. 0.4 A /
ÔÓËÁ‚Ó‰Ë
ÚÂθ: Friwo
ç‡ ‚˚ıӉ 5 VDC
íËÔ 15.2630
ìÒÎÓ‚Ëfl ˝ÍÒÔÎÛ‡Ú‡ˆËË
íÂÏÔÂ‡ÚÛ‡ ÓÚ +10 °C ‰Ó +35 °C
éÚÌÓÒËÚÂθ̇fl ‚·ÊÌÓÒÚ¸ ÓÚ 30% ‰Ó 75%
ч‚ÎÂÌË ÓÚ 700 hPa ‰Ó 1060 hPa
ê‡ÁÏÂ˚ Á‡fl‰ÌÓ„Ó ·ÎÓ͇ Ñ=205 ÏÏ; ò=150 ÏÏ; Ç=85 ÏÏ
ÇÂÒ Á‡fl‰ÌÓ„Ó ·ÎÓ͇ 250 „
ÇÂÏfl Á‡fl‰ÍË ‡ÍÍÛÏÛÎflÚÓ‡ ÔËÏ. 2 ˜ (ÔË ÔÓÎÌÓÒÚ¸˛ ‡ÁflÊÂÌÌÓÏ
‡ÍÍÛÏÛÎflÚÓÂ)
ùÎÂÍÚÓÔËÚ‡ÌËfl ̇ÍÓ̘ÌË͇ ‡ÍÍÛÏÛÎflÚÓ Li-Po
å‡ÍÒ.·Ó˜Â ÂÏflÍÍÛÏÛÎflÚÓ‡ ÔËÏ. 60 ÏËÌ. (ÔË ÌÓ‚Ó
Ï, ÔÓÎÌÓÒÚ¸˛
Á‡flÊÂÌÌÓÏ ‡ÍÍÛÏÛÎflÚÓÂ)
àÒÚÓ˜ÌËÍ Ò‚ÂÚ‡ ÔÓÎË‚ÓÎÌÓ‚ÓÈ ‰ËÓ‰
ÑˇԇÁÓÌ Ò‚ÂÚÓ‚ÓÈ ‚ÓÎÌ˚ 385–515 ÌÏ
åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl χÍÒ. 1.200 ÏÇÚ/ÒÏ
2
± 10%
êÂÊËÏ ‡·ÓÚ˚ 5 ÏËÌ. ‚Íβ˜ÂÌ / 6 ÏËÌ. ‚˚Íβ˜ÂÌ
(Ò Ô‡ÛÁ‡ÏË)
ê‡ÁÏÂ˚ ̇ÍÓ̘ÌË͇: Ñ=260 ÏÏ; ò=42 ÏÏ; Ç=120 ÏÏ
ÇÂÒ Ì‡ÍÓ̘ÌË͇ (c ‡ÍÍÛÏÛÎflÚÓÓÏ) 225
ìÒÎÓ‚Ëfl Ú‡ÌÒÔÓÚËÓ‚ÍË Ë ı‡ÌÂÌËfl:
íÂÏÔÂ‡ÚÛ‡ ÓÚ -20 °C ‰Ó +60 °C
éÚÌÓÒËÚÂθ̇fl ‚·ÊÌÓÒÚ¸ ÓÚ 10% ‰Ó 75%
ч‚ÎÂÌ
ËÂ ÓÚ 500 hPa ‰Ó 1060 hPa
ã‡ÏÔÛ Bluephase N ı‡ÌËÚ¸ ‚ Á‡Í˚ÚÓÏ ÔÓÏ¢ÂÌËË.
èË·Ó Ì ÔÓ‰‚Â„‡Ú¸ ÒËθÌ˚Ï ÒÓÚflÒÂÌËflÏ
ÄÍÍÛÏÛÎflÚÓ
– Ì ı‡ÌËÚ¸ ÔË ÚÂÏÔÂ‡ÚÛ ‚˚¯Â 40°C (ËÎË Í‡ÚÍÓ‚ÂÏÂÌÌÓ ÔË
60°C), ÂÍÓÏẨÛÂχfl ÚÂÏÔÂ‡ÚÛ‡ ı‡ÌÂÌËfl 15 – 30°C.
– ı‡ÌËÚ¸ ‚Ò„‰‡ Á‡flÊÂÌÌ˚Ï Ë
– Ì ‰Óθ¯Â 6 ÏÂÒflˆÂ‚
Bluephase-N_tr_ru_rev2_layout 06.06.16 13:42 Seite 28
86
Appendix
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment - guidance
The “Bluephase N” uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
The “Bluephase N” is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
P < 75W (pass without test)
P < 75W (pass without test)
Guidance and manufacturer‘s declaration - electromagnetic emission
The “Bluephase N” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase N” should
assure that it is used in such an environment.
Guidance and manufacturer‘s declaration - electromagnetic immunity
The “Bluephase N” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase N” should
assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line - line
± 2 kV line - earth
<5 % U
T
(>95 % dip in U
T
) for 0.5 cycle
40 % U
T
(60 % dip in U
T
) for 5 cycles
70 % U
T
(30 % dip in U
T
) for 25 cycles
<5 % U
T
(>95 % dip in U
T
) for 5 sec
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line - line
no prot. earth
<5 % U
T
(>95 % dip in U
T
) for 0.5 cycle
40 % U
T
(60 % dip in U
T
) for 5 cycles
70 % U
T
(30 % dip in U
T
) for 25 cycles
<5 % U
T
(>95 % dip in U
T
) for 5 sec
30 A/m
Electromagnetic environment - guidance
Floors should be concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Mains power quality should be that of typical commercial or dental environment.
Mains power quality should be that of typical commercial or dental environment.
Mains power quality should be that of typical commercial or dental environment. If the
user of the „Bluephase N“ requires continued operation during power mains interrup
-
tions, it is recommended that the „Bluephase N“ be powered from an uninterruptible
power supply or battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in
a typical commercial or dental environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
87
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 test level
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance level
10 V
10 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should not be used closer to any
part of the „Bluephase N“, including cables, than the recommended separation distance
caculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 0.35
P
d = 0.35 P 80 MHz to 800 MHz
d = 0.70P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
Guidance and manufacturer‘s declaration - electromagnetic immunity
The “Bluephase N” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase N” should
assure that it is used in such an environment.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be con-
sidered. If the measured field strength in the location in which the „Bluephase N“ is used
exceeds the applicable RF compliance level above, the „Bluephase N“ should be observed
to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the „Bluephase N“.
b
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the “Bluephase N”
The “Bluephase N” is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the “Bluephase N” can help prevent electromagnetic interference by maintaining a miminum distance between portable
and mobile RF communications equipment (transmitters) and the „Bluephase N“ as recommended below,
according to the maximum output power of the communication equipment.
Rated
maximum
output power
of transmitter
W
0.01
0.1
1
10
100
150 kHz to
80 MHz
d = 0.4 P
0.04
0.13
0.40
1.3
4.0
Separation distance according to
frequency of transmitter
m
80 MHz to
800 MHz
d = 0.4 P
0.04
0.13
0.40
1.3
4.0
800 MHz to
2.5 GHz
d = 0.7 P
0.07
0.22
0.70
2.2
7.0
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Date information prepared: 2016-06 / Rev. 2
657717/2016-06-09
Ivoclar Vivadent – worldwide
Ivoclar Vivadent AG
Bendererstrasse 2
9494 Schaan
Liechtenstein
Tel. +423 235 35 35
Fax +423 235 33 60
www.ivoclarvivadent.com
Ivoclar Vivadent Pty. Ltd.
1 – 5 Overseas Drive
P.O. Box 367
Noble Park, Vic. 3174
Australia
Tel. +61 3 9795 9599
Fax +61 3 9795 9645
www.ivoclarvivadent.com.au
Ivoclar Vivadent GmbH
Tech Gate Vienna
Donau-City-Strasse 1
1220 Wien
Austria
Tel. +43 1 263 191 10
Fax: +43 1 263 191 111
www.ivoclarvivadent.at
Ivoclar Vivadent Ltda.
Alameda Caiapós, 723
Centro Empresarial Tamboré
CEP 06460-110 Barueri – SP
Brazil
Tel. +55 11 2424 7400
Fax +55 11 3466 0840
www.ivoclarvivadent.com.br
Ivoclar Vivadent Inc.
1-6600 Dixie Road
Mississauga, Ontario
L5T 2Y2
Canada
Tel. +1 905 670 8499
Fax +1 905 670 3102
www.ivoclarvivadent.us
Ivoclar Vivadent Shanghai
Trading Co., Ltd.
2/F Building 1, 881 Wuding Road,
Jing An District
200040 Shanghai
China
Tel. +86 21 6032 1657
Fax +86 21 6176 0968
www.ivoclarvivadent.com
Ivoclar Vivadent Marketing Ltd.
Calle 134 No. 7-B-83, Of. 520
Bogotá
Colombia
Tel. +57 1 627 3399
Fax +57 1 633 1663
www.ivoclarvivadent.co
Ivoclar Vivadent SAS
B.P. 118
F-74410 Saint-Jorioz
France
Tel. +33 4 50 88 64 00
Fax +33 4 50 68 91 52
www.ivoclarvivadent.fr
Ivoclar Vivadent GmbH
Dr. Adolf-Schneider-Str. 2
D-73479 Ellwangen, Jagst
Germany
Tel. +49 7961 889 0
Fax +49 7961 6326
www.ivoclarvivadent.de
Wieland Dental + Technik
GmbH & Co. KG
Lindenstrasse 2
75175 Pforzheim
Germany
Tel. +49 7231 3705 0
Fax +49 7231 3579 59
www.wieland-dental.com
Ivoclar Vivadent Marketing
(India) Pvt. Ltd.
503/504 Raheja Plaza
15 B Shah Industrial Estate
Veera Desai Road, Andheri (West)
Mumbai, 400 053
India
Tel. +91 22 2673 0302
Fax +91 22 2673 0301
www.ivoclarvivadent.in
Ivoclar Vivadent s.r.l.
Via Isonzo 67/69
40033 Casalecchio di Reno (BO)
Italy
Tel. +39 051 6113555
Fax +39 051 6113565
www.ivoclarvivadent.it
Ivoclar Vivadent K.K.
1-28-24-4F Hongo
Bunkyo-ku
Tokyo 113-0033
Japan
Tel. +81 3 6903 3535
Fax +81 3 5844 3657
www.ivoclarvivadent.jp
Ivoclar Vivadent Ltd.
12F W-Tower
54 Seocho-daero 77-gil, Seocho-gu
Seoul, 06611
Republic of Korea
Tel. +82 2 536 0714
Fax +82 2 596 0155
www.ivoclarvivadent.co.kr
Ivoclar Vivadent S.A. de C.V.
Calzada de Tlalpan 564,
Col Moderna, Del Benito Juárez
03810 México, D.F.
México
Tel. +52 55 50 62 10 00
Fax +52 55 50 62 10 29
www.ivoclarvivadent.com.mx
Ivoclar Vivadent BV
De Fruittuinen 32
2132 NZ Hoofddorp
Netherlands
Tel. +31 23 529 3791
Fax +31 23 555 4504
www.ivoclarvivadent.com
Ivoclar Vivadent Ltd.
12 Omega St, Rosedale
PO Box 303011 North Harbour
Auckland 0751
New Zealand
Tel. +64 9 914 9999
Fax +64 9 914 9990
www.ivoclarvivadent.co.nz
Ivoclar Vivadent Polska Sp. z o.o.
Al. Jana Pawla II 78
00-175 Warszawa
Poland
Tel. +48 22 635 5496
Fax +48 22 635 5469
www.ivoclarvivadent.pl
Ivoclar Vivadent Marketing Ltd.
Prospekt Andropova 18 korp. 6/
office 10-06
115432 Moscow
Russia
Tel. +7 499 418 0300
Fax +7 499 418 0310
www.ivoclarvivadent.ru
Ivoclar Vivadent Marketing Ltd.
Qlaya Main St.
Siricon Building No.14, 2
nd
Floor
Office No. 204
P.O. Box 300146
Riyadh 11372
Saudi Arabia
Tel. +966 11 293 8345
Fax +966 11 293 8344
www.ivoclarvivadent.com
Ivoclar Vivadent S.L.U.
Carretera de Fuencarral nº24
Portal 1 – Planta Baja
28108-Alcobendas (Madrid)
Spain
Telf. +34 91 375 78 20
Fax: +34 91 375 78 38
www.ivoclarvivadent.es
Ivoclar Vivadent AB
Dalvägen 14
S-169 56 Solna
Sweden
Tel. +46 8 514 939 30
Fax +46 8 514 939 40
www.ivoclarvivadent.se
Ivoclar Vivadent Liaison Office
: Tesvikiye Mahallesi
Sakayik Sokak
Nisantas’ Plaza No:38/2
Kat:5 Daire:24
34021 Sisli – Istanbul
Turkey
Tel. +90 212 343 0802
Fax +90 212 343 0842
www.ivoclarvivadent.com
Ivoclar Vivadent Limited
Ground Floor Compass Building
Feldspar Close
Warrens Business Park
Enderby
Leicester LE19 4SE
United Kingdom
Tel. +44 116 284 7880
Fax +44 116 284 7881
www.ivoclarvivadent.co.uk
Ivoclar Vivadent, Inc.
175 Pineview Drive
Amherst, N.Y. 14228
USA
Tel. +1 800 533 6825
Fax +1 716 691 2285
www.ivoclarvivadent.us

Transcripción de documentos

Bluephase N ® EN Instructions for Use Page 2 PT Instruções de Uso Página 44 FR Mode d’emploi Page 16 TR Kullanma Talimatı Sayfa 58 ES Instrucciones de uso Pagina 30 RU àÌÒÚÛ͈Ëfl ëÚ. 72 Rx ONLY For dental use only! Made in Austria Manufacturer: Ivoclar Vivadent AG FL-9494 Schaan/Liechtenstein In USA distributed by Ivoclar Vivadent Inc., 175 Pineview Drive, Amherst, NY 14228, USA Appendix Page 86 Dear Customer, Optimum polymerization is an important requirement for all light cured materials in order to consistently produce high quality restorations. The polymerization light selected also plays a decisive role in this respect. Therefore, we would like to thank you for having purchased Bluephase N. Bluephase N is a high-quality medical device which has been designed according to the latest standard of science and technology in compliance with the relevant industry standards. These Instructions for Use will help you safely start-up the device, make full use of its capabilities, and ensure a long service life. Should you have any further questions, please do not hesitate to contact us (see addresses on the reverse page). Your Ivoclar Vivadent Team 2 Product overview – List of parts  – Indicators on the charging base  – Indicators on the handpiece  English Table of Contents 4 Safety6 – Intended use  – Indication  – Signs and symbols – Safety notes Start-up  – Charging base  – Handpiece – Battery  – Click & Cure corded operation  8 Operation11 – Selecting the curing program and the curing time – Cure Memory function  – Start – Acoustic signals  – Light intensity  – Measuring the light intensity  Maintenance and cleaning – Housing – Light probes  – Battery contacts  – Disposal 13 What if …? 14 Warranty / Procedure in case of repair  15 Product specifications  15 3 List of parts 13 12 1 3 5 6 4 10 2 11 9 1 Handpiece 2 Start button 3 Display 4 Program selection button 5 Time selection button 6 Volume button 7 Power indicator 8 Charging base 9 Power cord 10 Power pack 11 Battery 12 Light probe 10 mm 13 Anti-glare cone 4 7 8 Indicators on the charging base Indicator lights up in blue - the charging base is connected to the power supply Indicators on the handpiece Curing program and operating state Selected indications and curing time Charging status of the battery 5 Safety Intended use Bluephase N is an LED polymerization light that produces energy-rich blue light. It is used for the polymerization of light-curing dental materials immediately at the dental unit. The intended use also includes the observation of the notes and regulations in these Instructions for Use. Safety notes The Bluephase N is is an electronic device and a medical product which is subject to IEC 60601-1 (EN 60601-1) and EMC directives, as well as the 93/42/EEC Medical Device Directive. The appliance complies with the ­relevant EU regulations and is classified as an LED Class 2 product. Indication With its “Polywave®” broadband spectrum, Bluephase N is suitable for the polymerization of all light-curing dental materials curing in the wavelength range of 385–515 nm. These materials include restoratives, bonding agents/ adhesives, bases, liners, fissure sealants, temporaries, as well as luting materials for brackets and dental-lab materials such as ceramic inlays. Signs and symbols Non permissible use Symbols on the curing light Double insulation (device complies with safety class II) Protection against electrical shock (BF type apparatus) Observe Instructions for Use Observe Instructions for Use Caution The curing light must not be disposed of in the normal domestic waste. Information ­regarding disposal of the light can be found on the respective national Ivoclar Vivadent homepage. Recyclable 6 The apparatus has been shipped from the manufacturer in a safe and technically sound condition. In order to maintain this condition and to ensure risk-free operation, the notes and regulations in these Instructions for Use have to be observed. To prevent damage to equipment and risks for patients, users, and third parties, the ­following safety instructions have to be observed. Materials, the polymerization of which is activated outside the wavelength range of 385–515 nm (no materials known to date.) If you are not sure about certain products, please ask the manufacturer of the corresponding material. Do not charge or use the appliance near ­flammable or combustible substances. Portable and mobile high-frequency communication devices may interfere with medical equipment. The use of mobile phones during operation is not allowed.                                                                                                        Usage and liability – Bluephase N must only be employed for the intended use. Any other uses are contraindicated. Liability cannot be accepted for damage resulting from misuse or failure to observe the Operating Instructions. – The user is responsible for testing Bluephase N for its use and suitability for the intended purposes. This is particularly important if other equipment is used in the immediate vicinity of Bluephase N at the same time. – Use only original spare parts and accessories from ­Ivoclar Vivadent (see Accessories). The manufacturer does not accept any liability for damage resulting from the use of other spare parts or accessories. – The light probe is an applied part and may warm up to a maximum of 45 °C (113 °F) at the interface to the handpiece during operation. Operating voltage Before switching on, make sure that a) the voltage indicated on the rating plate complies with the local power supply. b) the unit has acquired the ambient temperature. If the battery or power pack are used separately, e.g. during start-up or Click & Cure corded operation, contact with patients or third parties must be prevented. Do not touch the exposed contacts of the battery or connection plug (power pack). Assumption of impaired safety If it has to be assumed that safe operation is no longer possible, the power must be disconnected and the battery removed to avoid accidental operation. This may be the case, for example, if the apparatus is visibly damaged or no longer works correctly. A complete disconnection from the power supply is only ensured when the power cord is disconnected from the power source. Eye protection Direct or indirect exposure of the eyes must be prevented. Prolonged exposure to the light is unpleasant for the eyes and may result in injury. Therefore, using the supplied anti-glare cones is recommended. Individuals who are generally sensitive to light, who take photosensitizing drugs, have undergone eye surgery, or people who work with the apparatus or in its vicinity for long periods of time should not be exposed to the light of this device and wear protective goggles (orange) that absorb light below 515 nm. Battery Caution: Use only original spare parts, particularly Ivoclar Vivadent batteries and charging bases. Do not short circuit battery. Do not store at temperatures above 40 °C / 104 °F (or 60 °C / 140 °F for a short period). Always store batteries charged. The storage period must not exceed 6 months. May explode if disposed of in fire. Please note that the lithium polymer battery may react with explosion, fire, and smoke development in case of inappropriate handling and mechanical damage. Damaged lithium polymer batteries must no longer be used. The electrolytes and electrolyte fumes released during exlosion, fire, and smoke development are toxic and ­corrosive. In case of contact with the eyes and skin, immediately rinse with copious amounts of water. Avoid the inhalation of the fumes. See a physician immediately in case of indisposition. Heat development As it is the case with all high-performance lights, the high light intensity results in a certain heat development. Prolonged exposure of areas near the pulp and soft tissues may result in irreversible or reversible damage. Therefore, the recommended curing times have to be observed. Uninterrupted curing times of more than 20 seconds on the same tooth surface, as well as direct contact with the gingiva, oral mucous membrane, or skin, have to be prevented. If required, polymerize ­indirect restorations at intermitting intervals of 20 seconds each or use external cooling with an air stream. 7 Start-up Check the delivery for completeness and any possible transportation damage (see delivery form). If components are missing or damaged, immediately contact your dealer or your Service Center. Handpiece Before you insert the light probe, remove the protective foil on the corresponding opening of the handpiece. Charging base Before you switch on the device, make sure that the voltage mentioned on the rating plate complies with your local power supply. The rating plate is attached to the underside of the charging base. Slightly turn the light probe while attaching it to the handpiece. After that, mount anti-glare protection on the light probe. Place the charging base on a suitable, flat table top. Remove the protective foil from the connection plug of the power pack. Slide the connection plug of the power pack into the socket on the underside of the charging base. Tilt it slightly and apply slight pressure until you hear and feel it snap into place. Connect the power cord with the power supply and the power pack. The Power ON indicator on the left side of the housing lights up in blue (see Indicators on the charging base). 8 Battery We recommend fully charging the battery before the first use. If the battery is fully charged, it features a curing capacity of approximately 60 minutes. Charging status of the battery With the handpiece switched on, the current charging status is shown on the display as follows: Battery fully charged (curing capacity of approximately 60 minutes) Battery half full Reserve (The last bar in the battery indicator is red in the reserve mode. The battery has to be charged as soon as possible.) Slide the battery straight into the handpiece until you hear and feel it click into place. When the battery is being charged, „Charging Battery“ briefly appears in the display before the display returns to the stand-by mode. If the battery is completely empty, the handpiece automatically switches to the stand-by mode. The light can no longer be called up and the curing program and curing time can no longer be set. However, the handpiece can be used in the Click & Cure corded operation. Gently place the handpiece in the corresponding rest in the charging base. If a sleeve is used, please remove it before you charge the battery. If possible, use the light always with a fully charged battery. This will prolong the service life. It is therefore recommended to place the handpiece into the charging base after each patient. The charging time for empty batteries is 2 hours. Since the battery is a consumable, it has to be replaced after its typical life cycle has expired after 2½ years. See battery label for the age of the battery. 9 Operation Click & Cure corded operation Bluephase N can be used in corded operation at any time, but particularly when the battery is completely empty. For this purpose, remove the battery from the handpiece. Then remove the power pack from the underside of the charging base. Do not pull on the power cord. Disinfect or autoclave contaminated surfaces of the curing light, as well as light probes, and anti-glare cones before each use. Furthermore, make sure that the stipulated light intensity permits adequate polymerization. For that purpose, check the light probe for contamination and damage, as well as the light intensity at regular intervals (e.g. with the Ivoclar Vivadent radiometer “Bluephase Meter”). Selecting the curing program and time The curing program and the curing time can be individually set. Bluephase N is equipped with the ­following 3 curing programs for the different indications. The desired curing program is selected with the program selection buttons. The display changes accordingly (see Indicators on the handpiece). Upon the first start-up, the following settings have been preset: – HIGH (High Power): 10 seconds – LOW (Low Power): 10 seconds – SOFT (Soft start): 15 seconds The intended curing time is selected using the time selection buttons. Users may choose between 5, 10, 15, 20 and 30 seconds. Insert the connection plug straight into the handpiece until you hear and feel it click into place. During corded operation, the charging base cannot charge the battery. The apparatus is only completely disconnected from the power supply if the power cord has been pulled out of the plug-in socket. 10 Observe the Instructions for Use of the material applied when selecting the curing time. The curing recommendations for composite materials apply to all shades and, if not mentioned otherwise in the Instructions for Use of the relevant material, to a maximum layer thickness of 2 mm. Generally, these ­recommendations apply to situations where the emission window of the light probe is placed directly over the material to be polymerized. Increasing the distance between the light source and the material will require the curing time to be extended accordingly. For instance, if the distance to the material is 9 mm, the effective light output is reduced by approx. 50%. In this case, the recommended curing time has to be doubled. Curing programs HIGH POWER Program Consistently high light intensity for the polymerization of restorative and cementation materials for direct and indirect restorations. 1.200 mW/cm2 0 t [s] LOW POWER Program Reduced light intensity with reduced heat development for the polymerization of adhesives, liners, and restorative materials in areas near the pulp when restoring Class V cavities. 650 mW/cm2 0 1,200 mW/cm2 15 s 10 s Filling materials Composite • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow All conventional composites • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Compomer 3) Compoglass F / Compoglass Flow Indirect restorations / Luting materials Variolink II Base 4) / Variolink Veneer / Variolink N Base 4) / Variolink N Clear Veneer / Variolink Ultra 5) / Dual Cement 5) / Variolink II 5) Miscellaneous Helioseal / Helioseal F / Helioseal Clear Monopaque MultiCore Flow / Multicore HB Heliosit Orthodontic Telio Add-On Flow Telio Stains IPS Empress Direct Color IPS Empress Direct Opaque Curing time 10 s 15 s 10 s 20 s per mm ceramic: 10 s per segment 10 s 20 s 20 s 10 s 15 s 10 s 10 s 20 s Adhesive AdheSE / AdheSE One F ExciTE / ExciTE DSC Tetric N-Bond / Tetric N-Bond Self-Etch Heliobond (Syntac) Temporary materials Telio CS Inlay/Onlay Systemp.inlay/onlay Fermit / Fermit N Telio CS Link / Systemp.link Miscellaneous Heliosit Orthodontic Vivaglass Liner Curing time Filling materials Composite • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow All conventional composites • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Compomer 3) Compoglass F / Compoglass Flow Curing time 10 s 10 s 10 s 10 s 20 s per segment 20 s 20 s t [s] SOFT START Program Step-by-step increase of the light intensity with reduced shrinkage stress and reduced heat development for the polymerization of restorative materials. 1.200 mW/cm2 650 mW/cm 2 0 Light intensity Exposure time for composites Exposure time for Tetric EvoCeram / IPS Empress Direct 5 15 s 20 s 15 s 20 s t [s] 1) Applies to a maximum layer thickness of 2 mm and provided that the Instructions for Use of the respective material do not state any other r­ecommendation (might be the case, e.g. with dentin shades) 2) Applies to a maximum layer thickness of 4 mm and provided that the Instructions for Use of the respective material do not state any other r­ecommendation (might be the case, e.g. with dentin shades) 3) Applies to a maximum layer thickness of 3 mm 4) Applies to light-curing (use of the base past only) 5) Applies to dual-curing 11 Cure Memory function The last settings used, together with the combination of curing program and curing time, are automatically saved. Start The light is switched on by means of the start button. Once the selected curing time has elapsed, the curing program is automatically terminated. If desired, the light can be switched off before the set curing time has elapsed by pressing the start button again. The fan is activated simultaneously to the light. Once the curing time has elapsed, the fan continues to run for a certain time to cool the apparatus. The battery must not be removed as long as the fan is still running. Acoustic signals Acoustic signals can be heard for the following functions: – Start (Stop) – Every 10 seconds – Program change – Curing time change – Connecting the battery to the charging base – Inserting battery – Error message If desired, the volume of the acoustic signals can be adjusted. For that purpose, press the blue volume button while the light is off and reduce the volume or turn the acoustic signals off with the left program selection button. If you want to turn the acoustic signals back on or increase the volume, press the blue volume button and then the right program selection button. 12 Light intensity The light intensity is maintained at a consistent level during operation. If the supplied 10 mm light probe is used, the light intensity has been calibrated to 1200 mW/cm2 ± 10%. T he use of a light probe other than the one provided has a direct influence on the light intensity emitted. In parallel-walled light probes (10 mm), the diameter is equal at both ends. In focussing light probes (10>8 mm light probe, Pin-Point light probe 6>2 mm), the diameter of the rear end is larger than that of the light emission window. The incident blue light is thus bundled to a smaller surface, which increases the light intensity emitted. Pin-Point light probes are suitable for the polymerization of confined areas, such as the attachment of veneers prior to excess removal. For thorough curing, it is necessary to change the light probe. Measuring the light intensity The light intensity of the Bluephase N and the enclosed 10 mm light probe can be checked by means of the Bluephase Meter. If the measured value does not correspond with the expected light intensity, proceed as follows: – Check the selected curing program. – Clean possibly dirty light sensor (radiometer). – Remove the light probe and clean the light emission window of the handpiece with a cotton swab dipped in alcohol. – Clean possibly dirty light probe (see Maintenance and cleaning). – Replace a damaged light probe with a new one. If these measures do not improve the results, please contact your dealer or your local Service Center. Maintenance and cleaning For reasons of hygiene, we recommend using a disposable protective sleeve for each pa-tient. Make sure to fit the protective sleeve closely to the light probe. Disinfect ­contaminated surfaces of the device and anti-glare cones (FD 366/Dürr Dental, Incidin Liquid/Ecolab) and sterilize the light probe before each use if disposable protective sleeves are not used. Make sure that no liquids or other foreign substances enter the handpiece, charging base and particularly the power pack during cleaning (risk of electrical shock). Disconnect the charging base from the power source when cleaning it. Housing Wipe the handpiece and handpiece holder with a ­customary aldehyde-free disinfecting solution. Do not clean with highly aggressive disinfecting solutions (e.g. solutions based on orange oil or with an ethanol content of more than 40%), solvents (e.g. acetone), or pointed instruments, which may damage or scratch the plastic. Clean dirty plastic parts with a soapy solution. Light probe Before cleaning and/or disinfecting the light probe, ­pretreat it. This applies to both automated and manual cleaning and disinfection. Pretreatment – Remove gross contamination immediately after use or within 2 hours. For this purpose, thoroughly rinse the light probe under running water (for at least 10 seconds). Alternatively, use a suitable aldehyde-free disinfecting solution to prevent the fixation of blood. – To remove contamination manually, use a soft brush or soft cloth. Partially polymerized composite can be removed with alcohol and a plastic spatula, if necessary. Do not use sharp or pointed objects, as they may scratch the surface. Cleaning and disinfection For cleaning, immerse the light probe in a cleaning solution and ensure that it is sufficiently covered with liquid (ultrasound or careful brushing with a soft brush may ­support the effect). A neutral-enzymatic cleaning agent is recommended. When cleaning and disinfecting, please make sure that the agents used are free of: – organic, mineral and oxidizing acids (the minimum admissible pH value is 5.5) – alkaline solution (the maximum admissible pH value is 8.5) – oxidizing agent (e.g. hydrogen peroxide) Afterwards, remove the light probe from the solution and thoroughly rinse it under running water (for at least 10 seconds). Cleaning in a thermal disinfector is an ­effective alternative. Sterilization Thorough cleaning and disinfecting is imperative to ensure that the subsequent sterilization is effective. Use only autoclave sterilization for this purpose. The sterilization time (exposure time at sterilization temperature) is 4 minutes at 134 °C (273 °F); pressure should be 2 bar (29 psi). Dry the sterilized light probe using either the special drying program of your steam autoclave or hot air. The light probe has been tested for up to 200 sterilization cycles. After that, check the light probe for damage. Hold it against light. If individual segments appear black, glass fibres are broken. If this is the case, replace the light probe with a new one. Battery contacts To ensure reliable conductivity at all times, keep the ­battery contacts free from possible contamination (e.g. composite residue). For this purpose, clean the affected contacts regularly in the course of the usual wipe ­disinfection (after each patient). Disposal The curing light must not be disposed of as urban waste. Dispose unserviceable batteries and polymerization lights according to the corresponding legal requirements in your country. Batteries must not be incinerated. 13 What if....? Symbol No display during charging Causes Error Rectification Electronic defect in the handpiece or battery Remove and reinsert the battery. If the error remains, replace the battery with the power pack (Click & Cure). If the error remains, please contact your dealer or your local Service Center. Apparatus is overheated or undercooled Allow the apparatus to cool down (or assume room temperature if it is undercooled) and try again after a certain time. If the error remains, please contact your dealer or your local Service Center. Electronic defect in the battery Remove and reinsert the battery. If the error remains, place the apparatus into the charging base. If the error remains nonetheless, replace the battery with the power pack (Click & Cure). Please contact your dealer or your local Service Center. Battery empty Position the handpiece in the charging base. „Charging Battery“ briefly appears in the display before the display returns to the stand-by mode. If the battery is not charged, the contacts have to be cleaned. If the error remains, please contact your dealer or your local Service Center. Charging contacts contaminated Check if the handpiece has been correctly placed into the charging base. Clean battery contacts. Information: “Charging Battery“ briefly appears in the display before the display returns to the stand-by mode. Battery in safety mode (integrated protective circuit) If cleaning the battery contacts does not improve the results, remove the battery from the apparatus. Charge the battery separately from the handpiece in the charging base for approx. 10 minutes. If the error remains, please contact your dealer or your local Service Center. LED of the charging base is not illuminated 14 Power pack not connected or defective Check if the power pack is correctly positioned in the charging base (also check the charging contacts) or if the power pack is connected to the power supply by means of the power cord (display on the power pack lights up in green if it works correctly). If the error remains, please contact your dealer or your local Service Center. Warranty / Procedure in case of repair The warranty period for Bluephase N is 3 years from the date of purchase (battery 1 year). Malfunctions resulting from faulty material or manufacturing errors are repaired free of charge during the warranty period. The warranty does not provide the right to recover any material or non-material damage other than the ones mentioned. The apparatus must only be used for the intended purposes. Any other uses are contraindicated. The manufacturer does not accept any liability resulting from misuse and warranty claims ­cannot be accepted in such cases. This is particularly valid for: – damage resulting from improper handling, especially incorrectly stored batteries (see Technical data: Transportation and storage conditions). – damage to components resulting from wear under standard operating conditions (e.g. battery). – damage resulting from external influences, e.g. blows, drop to the floor. – damage resulting from incorrect set-up or installation. – damage resulting from connecting the unit to a power supply, the voltage and frequency of which do not comply with the ones stated on the rating plate. – damage resulting from improper repairs or modifications that have not been carried out by certified Service Centers. In case of a claim under warranty, the complete apparatus (handpiece, charging base, power cord, and power pack) must be returned, carriage paid, to the dealer or directly to Ivoclar Vivadent, together with the purchase document. Use the original packaging with the corresponding cardboard inserts for transportation. Repair work may only be carried out by a certified Ivoclar Vivadent Service Center. In case of a defect that cannot be rectified, please contact your dealer or your local Service Center (see addresses on the reverse side). A clear description of the defect or the conditions under which the defect occurred will facilitate locating the problem. Please enclose this description when returning the apparatus. Product specifications Delivery form 1 Charging base, 1 Power cord, 1 Power pack, 1 Handpiece, 1 Battery (Li-Po), 1 10 mm light probe, 3 Anti-glare cones, 1 Set of Instructions for Use Accessories The following accessories are available for Bluephase N: REF 656196 608538 551756 592496 659956 659681 607922 608554 Description Light probe 10 mm, black (N) Pin-point light probe 6>2 mm, black Anti-glare cones Anti-glare shield Bluephase N battery Bluephase N handpiece Bluephase Meter Bluephase sleeves Technical data Operating voltage charging base 5 VDC Operating voltage handpiece 3.7 VDC with battery 5 VDC with power pack Power pack 100–240 VAC / 50-60 Hz / max. 0.4 A Manufacturer: Friwo Output 5 VDC TYPE 15.2630 Operating conditions Temperature +10 °C to +35 °C (+50 to +95 °F) Relative humidity 30% to 75% Ambient pressure 700 hPa to 1060 hPa Dimensions of the charging base L=205mm W=150mm H=85mm Weight of the charging base 250 g Charging time approx. 2 h (with the battery empty) Power supply of the handpiece Li-Po battery Max. battery time approx. 60 min. (with a new, fully charged battery) Light source Polywave® LED Wavelength range 385 – 515 nm Light intensity max. 1,200 mW/cm2 ± 10 % Operating 5 min. on / 6 min. off (intermittently) Dimensions of the handpiece: L=260mm W=42mm H=120 mm Weight of the handpiece (incl. battery) 225 g Transportation and storage conditions: Temperature -20 °C to +60 °C (-4 to 140 °F) Relative humidity 10% to 75% Ambient pressure 500 hPa to 1060 hPa Bluephase N has to be stored in closed, roofed rooms. Protect the device from severe jarring. Battery – Do not store at temperatures above 40 °C / 104 °F (or 60 °C / 140 °F for a short period). Recommended storage temperature 15 – 30 °C (59 – 86 °F). – Always store the battery charged – and not for longer than 6 months. 15 Cher client, Une polymérisation optimale est une condition importante pour tous les matériaux photopoly­ mérisables de manière à produire de façon constante et reproductible des restaurations de haute qualité. La lampe à polymériser choisie a, elle aussi, un rôle décisif de ce point de vue. C‘est pour cela que nous tenons à vous remercier d‘avoir acheté la lampe Bluephase N. La lampe Bluephase N est un dispositif médical de haute qualité correspondant aux dernières normes scientifiques et technologiques ainsi qu‘aux normes industrielles. Ce mode d‘emploi va vous permettre de mettre en marche votre appareil en toute sécurité, de profiter de toutes ses possibilités, et de lui assurer une longue durée de vie. Pour une identification facile des composants individuels, la vue générale du produit est présentée sous la forme d‘une carte. Pour toute question supplémentaire, nous vous remercions de nous contacter (adresse sur le verso). Votre équipe Ivoclar Vivadent 16 Vue générale  – Listes des pièces  – Indicateurs sur la base de chargement  – Indicateurs sur la pièce à main  Francais Sommaire 18 Sécurité20 – Usage habituel  – Indication  – Signes et symboles  – Notes de sécurité Première mise en marche  – Base de chargement  – Pièce à main  – Batterie  – Cordon Click & Cure 22 Utilisation  – Sélection du programme et du temps de polymérisation – Fonction de mémoire de programme de polymérisation – Démarrage  – Signaux acoustiques – Intensité lumineuse – Mesurer l’intensité lumineuse 24 Maintenance et nettoyage  – Logement – Conducteurs de lumiére – Contacts de la batterie  – Elimination des pièces usagées 27 Que faire en cas de …?  28 Garantie / Procédure en cas de réparation  29 Spécifications  29 17 Liste des pièces détachées 13 12 1 3 5 6 4 10 2 11 9 1 Pièce à main 2 Bouton de mise marche 3 Affichage 4 Bouton de sélection des programmes 5 Bouton de sélection du temps 6 Bouton de réglage du volume 7 Indicateur de puissance 8 Base de chargement 9 Cordon d‘alimentation 10 Power pack 11 Batterie 12 Conducteur de lumière 10 mm 13 Embout de protection oculaire 18 7 8 Indicateurs sur la base de chargement Les voyants s‘allument en bleu – la base de chargement est reliée au courant électrique. Indicateurs sur la pièce à main Programme de polymérisation, temps de polymérisation, état de marche Indications spécifiques du programme de polymérisation sélectionné Témoin de charge de la batterie 19 Sécurité Usage habituel Bluephase N est une lampe à photopolymériser LED de haute performance qui produit une lumière bleue dense de haute énergie. Elle est utilisée pour la polymérisation de matériaux dentaires au fauteuil. L‘utilisation requise inclut aussi l‘observation des recommandations et réglementations du mode d‘emploi. Notes de sécurité Bluephase N est un appareil médical électronique ­fabriqué conformément à la norme IEC 60601-1 (EN 60601-1 et répondant aux Directives CEM, ainsi qu‘aux Directives des Dispositifs Médicaux 93/42/EEC. L’appareil est conforme aux normes européennes en vigueur et il est ­répertorié comme un produit LED de Classe 2. Indications Grâce à son spectre à large longueur d‘ondes, Bluephase N est particulièrement recommandée pour la polymérisation de matériaux dentaires photopolymérisables dans une longueur d‘onde de 385–515 nm. Ces matériaux incluent les matériaux de restauration, les bondings/ adhésifs, les bases, les liners, les scellements de sillons et fissures, les matériaux provisoires ainsi que les colles pour brackets orthodontiques, les matériaux de laboratoire dentaire tels que les inlays céramiques. Signes et symboles Contre-indication Symboles sur la lampe à photopolymériser Double protection (l‘appareil répond aux normes de sécurité de classe II) Protection contre les décharges électriques (type d‘appareil BF) Respecter le mode d‘emploi Respecter le mode d‘emploi Mise en garde Ne pas jeter la lampe avec les ordures ménagères. Concernant l‘élimintation de votre appareil, veuillez contacter votre ­distributeur habituel de produits Ivoclar Vivadent. Recyclable 20 La lampe est expédiée par le fabricant dans des ­conditions de sécurité et des conditions techniques ­optimales. Pour maintenir ces conditions et assurer des opérations sans risque, il est nécessaire de respecter les recommandations et les réglementations du mode ­d‘emploi. Pour prévenir les dommages sur l‘appareil et tout risque pour les patients, les utilisateurs et les tiers, les instructions de sécurité suivantes doivent être respectées. Ne pas utiliser l‘appareil pour photopolymériser les matériaux dont la polymérisation est activée dans une longueur d‘onde en dehors de 385–515 nm (aucun matériau connu à ce jour). Si vous avez un doute sur certains produits, contacter le fabricant du matériau. Ne pas charger ou utiliser cet appareil à ­proximité de substances inflammables ou combustibles. Les téléphones portables et autres appareils de communication HF (Haute Fréquence) peuvent interférer les équipements médicaux. L‘utilisation d‘un téléphone portable pendant le fonctionnement de la lampe est interdite. Manipulation et responsabilité – Bluephase N ne doit être utilisée que pour les usages recommandés. Toute autre utilisation est contre-­ indiquée. Nous rejetons toute responsabilité pour des dommages liés à un usage inadapté ou une panne pour non-respect du présent mode d‘emploi. – De plus, l‘utilisateur est responsable de l‘usage de la lampe Bluephase N de façon appropriée en respectant les indications recommandées, et ce particulièrement si d‘autres équipements sont utilisés à proximité ­pendant le fonctionnement de la lampe. – N‘utiliser que les pièces détachées et accessoires ­fournis par Ivoclar Vivadent (voir Accessoires). Nous ­rejetons toute responsabilité pour les dommages ­causés par l‘utilisation d‘autres pièces détachées ou accessoires. – L’embout lumineux est une pièce appliquée et, ­pendant le fonctionnement, peut chauffer jusqu’à 45 °C m ­ aximum au niveau de l’interface avec la pièce à main. recommandée. Les personnes sensibles à la lumière, qui prennent des médicaments photosensibles ou qui ont subi une chirurgie oculaire, les personnes qui travaillent avec l‘appareil ou à proximité pendant de longues périodes ne doivent pas être exposées à la lumière de cette lampe et doivent porter des lunettes de protection qui absorbent la lumière en dessous de 515 nm. Batterie Attention : N‘utiliser que les pièces d‘origine. En particulier, n‘utiliser que les batteries et bases de chargement ­d‘origine Ivoclar Vivadent. Ne pas mettre la batterie en court-circuit. Ne pas stocker la batterie à une température supérieure à 60°C, ni pendant plus de 6 mois sans être rechargée. Ne pas brûler (risque d‘explosion). Voltage Avant la mise en marche, assurez-vous que : a) le voltage indiqué sur la base est conforme à l‘énergie délivrée par l‘alimentation électrique. et b) l‘appareil se trouve à température ambiante. Notez que toute flamme, explosion ou dégagement de fumée dus à une mauvaise manipulation ou à un problème mécanique peut provoquer une réaction de la batterie Lithium Polymère. Les batteries Lithium Polymère endommagées ne doivent plus être utilisées. Si la batterie ou le cordon d‘alimentation sont utilisés séparément, par exemple lors de la mise en marche ou lors de l‘opération Click & Cure raccordée au courant, les contacts avec les patients ou les tiers doivent être évités. Ne pas toucher les contacts accessibles de la batterie ou les connexions électriques (power pack). Les electrolytes et les fumées d‘électrolytes libérés par une explosion, une flamme ou un dégagement de fumée sont toxiques et corrosifs. En cas de contact avec les yeux ou la peau, rincer immédiatement et abondamment à l‘eau. Eviter d‘inhaler les fumées. Consulter rapidement votre médecin en cas d‘indisposition. Conditions de sécurité Si un usage en toute sécurité ne peut être garanti, la prise de courant doit être débranchée et la batterie retirée pour éviter tout fonctionnement accidentel. Ceci peut être le cas, si l‘appareil est visiblement endommagé ou ne fonctionne plus correctement. Ne pas toucher les connexions exposées des batteries ou du cordon d‘alimentation. La déconnexion complète n‘est assurée que lorsque le cordon d‘alimentation est débranché. Dégagement de chaleur Comme cela est le cas avec toutes les lampes à haute performance, une haute énergie lumineuse a pour résultat un certain dégagement de chaleur. Une exposition prolongée sur la pulpe ou sur les tissus mous peut engendrer des dommages. Il est donc conseillé de respecter les recommandations de polymérisation liées au programme et au temps de polymérisation. De plus, une polymérisation continue de plus de 20 secondes sur la même zone, ainsi qu‘un contact direct sur la gencive, les muqueuses, ou la peau, doivent être évités. Si nécessaire, polymériser les restaurations indirectes par intervalles de 20 secondes ou utiliser un jet d‘air qui permettra le refroidissement. Protection oculaire Eviter toute exposition directe ou indirecte avec les yeux. Une exposition prolongée à la lumière peut être inconfortable et provoquer des dommages. C‘est pour cela, que l‘utilisation des embouts de protection oculaire est 21 Mise en marche Contrôler la livraison afin de vérifier qu‘il ne manque rien et qu‘aucun dommage n‘est survenu lors du transport (voir livraison). Si des composants manquent ou sont endommagés, contacter immédiatement votre revendeur ou centre de service. Pièce à main Avant d‘insérer la fibre de verre conductrice de lumière, retirer le film de protection sur l‘ouverture de la pièce à main. Base de chargement Avant de mettre en marche l‘appareil, assurez-vous que le voltage inscrit sur la base de chargement est conforme à votre alimentation électrique. Les indications sont portées sur la plaque d‘identification qui se trouve sous la base de chargement. Tourner légèrement le conducteur de lumière dès que celui-ci est fixé sur la pièce à main, puis positionner l‘embout de protection oculaire sur le conducteur de lumière. Poser la base de chargement sur une surface plate et ­stable. Mettre en place la prise de connexion au niveau du compartiment sous la base de chargement. L‘incliner légèrement et exercer une légère pression jusqu‘à sentir et entendre un clic. Connecter le cordon d‘alimentation à la prise de courant. L‘indicateur Power On sur le côté gauche du capot de la base de chargement s‘allume en vert (voir “Indicateurs sur la base de chargement”). 22 Batterie Il est conseillé d‘effectuer une charge complète de la batterie avant la première utilisation. Une batterie complètement chargée a une capacité de polymérisation d‘environ 60 minutes. Etat de charge de la batterie Lorsque la pièce à main est allumée, l‘état de la charge est affiché sur l‘écran d‘affichage comme suit (voir “Indicateurs sur la pièce à main”) : Batterie complètement chargée (capacité de polymérisation d‘environ 60 minutes) Batterie à moitié chargée Faire glisser la batterie dans la pièce à main jusqu‘à l‘obtention du clic de mise en place. Réserve (la dernière barre de l‘indicateur de la batterie devient rouge en mode reserve. La batterie doit être mise en charge dès que possible). Lorsque la batterie est mise en charge, „Charging Battery“ s‘affiche brièvement avant le retour en mode stand-by de l‘écran. Reposer délicatement la pièce à main sur la base de chargement, à l‘emplacement prévu à cet effet. Si vous utilisez une housse de protection, n‘oubliez pas de la retirer avant de mettre la batterie en charge. Si possible, toujours utiliser la lampe avec une batterie complètement chargée, afin d‘en prolonger la durée de vie. Il est conseillé de replacer la pièce à main sur la base de chargement après chaque patient. Le temps de charge d‘une batterie vide est de 2 heures. Batterie vide Si la batterie est complètement vide, la pièce à main se met automatiquement en mode stand-by. La lumière ne s‘allume plus, le programme de polymérisation et le temps de polymérisation ne peuvent être réglés. Cependant, la pièce à main peut être utilisée avec le cordon Click & Cure. La batterie est un consommable et doit, de ce fait, être remplacée environ tous les 2½ ans, cycle de vie d‘une batterie. Voir l‘âge de la batterie sur l‘étiquette. 23 Utilisation Cordon Click & Cure La pièce à main Bluephase N peut être branchée au courant à n‘importe quel moment, plus particulièrement en cas d‘urgence alors que la batterie est vide. Pour cela, retirer la batterie de la pièce à main. Puis ­retirer le cordon de branchement „power pack“ de son logement sous le socle de la base de chargement. Ne pas tirer sur le cordon. Insérer le cordon de branchement “power pack” dans la pièce à main jusqu‘à sentir et entendre un clic. Pendant que la pièce à main est connectée au courant, la base de chargement ne peut charger la batterie. La déconnexion complète n‘est assurée que lorsque le cordon d‘alimentation est débranché. 24 Désinfecter ou passer à l‘autoclave les différentes sur­ faces de la lampe ainsi que les conducteurs de lumière et l‘embout de protection oculaire avant chaque utilisation. Par ailleurs, assurez-vous que l‘intensité lumineuse obtenue permet une polymérisation adéquate. Pour cela, vérifier que le conducteur de lumière n‘est ni souillé ni endommagé, puis contrôler l‘intensité régulièrement (par ex. avec le radiomètre “Bluephase Meter” d‘Ivoclar Vivadent). Sélection du programme et du temps de polymérisation Le programme et le temps de polymérisation peuvent être programmés individuellement. Bluephase N est équipée des 3 programmes suivants pour les différentes indications. Le programme de polymérisation souhaité est sélectionné à l‘aide du bouton de sélection du ­programme. L‘affichage change en conséquence (voir “Indicateurs sur la pièce à main”). Les programmes suivants sont pré-enregistrés : – HIGH (Haute énergie) : 10 secondes – LOW (Basse énergie) : 10 secondes – SOFT (Polymérisation progressive) : 15 secondes Le temps de polymérisation souhaité est sélectionné à l‘aide du bouton de sélection du temps. Les utilisateurs ont le choix entre 5, 10, 15, 20 et 30 secondes. Respecter le mode d‘emploi des matériaux utilisés lors de la sélection du programme. Les recommandations de polymérisation des matériaux composites s‘appliquent à toutes les teintes. Sauf indication contraire dans le mode d‘emploi du matériau utilisé, photopolymériser par couche de 2 mm maximum. Généralement, ces recommandations s‘appliquent dans les situations où le conducteur de lumière est orienté directement sur le matériau à polymériser. Augmenter la distance entre la source de lumière et le matériau demande d‘augmenter le temps de polymérisation en conséquence. Par exemple, si la distance avec le ­matériau est de 5 mm, l‘intensité lumineuse réelle est réduite d‘environ 50%. Dans ce cas, il convient de ­doubler le temps de polymérisation. Programmes de polymérisation HIGH POWER (Haute énergie) Haute énergie régulière pour la polymérisation de matériaux de restauration et de collage dans le cas de restaurations directes et indirectes. 1.200 mW/cm 2 0 t [s] 650 mW/cm2 0 LOW POWER Basse énergie, élévation de température réduite pour la polymérisation des adhésifs, des liners et des matériaux de restauration pour les cavités de classe V dans les zones proches de la pulpe. Matériaux de restaurations Composite • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow tous les composites conventionnels • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Compomère 3) Compoglass F / Compoglass Flow Restaurations indirectes / Colles Variolink II Base 4) / Variolink Veneer / Variolink N Base 4) / Variolink N Clear Veneer / Variolink Ultra 5) / Dual Cement 5) / Variolink II 5) Divers Helioseal / Helioseal F / Helioseal Clear Monopaque MultiCore Flow / Multicore HB Heliosit Orthodontic Telio Add-On Flow Telio Stains IPS Empress Direct Color IPS Empress Direct Opaque Temps de polymérisation 10 secondes 15 secondes 10 secondes 20 secondes Par mm de céramique : 10 secondes par face 10 secondes 20 secondes 20 secondes 10 secondes 15 secondes 10 secondes 10 secondes 20 secondes Adhesifs AdheSE / AdheSE One F ExciTE / ExciTE DSC Tetric N-Bond / Tetric N-Bond Self-Etch Heliobond (Syntac) Matériaux temporaires Telio CS Inlay/Onlay Systemp.inlay/onlay Fermit / Fermit N Telio CS Link / Systemp.link Divers Heliosit Orthodontic Vivaglass Liner Temps de polymérisation Matériaux de restaurations Composite • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow tous les composites conventionnels • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Compomére 3) Compoglass F / Compoglass Flow Temps de polymérisation 10 secondes 10 secondes 10 secondes 10 secondes 20 secondes par face 20 secondes 20 secondes t [s] 1.200 mW/cm2 SOFT START Augmentation progressive de l‘intensité lumineuse avec une réduction du stress de polymérisation et de la température pour la polymérisation des matériaux de restauration. 650 mW/cm2 0 Intensité 1'200 mW/cm2 15 secondes Temps d‘exposition pour les composites Temps d‘exposition pour Tetric EvoCeram / IPS Empress Direct 10 secondes 5 15 secondes 20 secondes ­ 15 secondes 20 secondes t [s] 1) ­Valable pour une épaisseur de 2 mm maximum et dans la mesure où le mode d‘emploi du matériau utilisé n‘indique pas d‘autre recommandation (comme cela peut être le cas par exemple avec les teintes Dentine) 2) Valable pour une épaisseur de 4 mm maximum et dans la mesure où le mode d‘emploi du matériau utilisé n‘indique pas d‘autre recommandation (comme cela peut être le cas par exemple avec les teintes Dentine) 3) Valable pour une épaisseur de 3 mm maximum 4) Valable pour la photopolymérisation (utilisation de la pâte base uniquement) 5) Valable pour la polymérisation duale 25 Fonction mémoire de polymérisation Les derniers réglages utilisés, programme de polymérisation et temps de polymérisation, sont automatiquement mémorisés. Démarrage La lampe est allumée avec le bouton de démarrage. Une fois le temps de polymérisation écoulé, le programme de polymérisation s‘arrête automatiquement. La polymérisation peut être interrompue avant la fin du temps complet en appuyant sur le bouton de démarrage. Le ventilateur se déclenche en même temps que la lumière. Une fois la polymérisation terminée, le ventilateur continue à ­fonctionner un certain temps pour refroidir la lampe. La batterie ne doit pas être retirée tant que le ventilateur continue à fonctionner. Signaux acoustiques Des signaux acoustiques sont émis lors des fonctions suivantes : – Démarrage (Arrêt) – Toutes les 10 secondes – Changement de programme – Changement du temps de polymérisation – Connexion de la batterie à la base de chargement – Insertion de la batterie – Message d‘erreur Le volume des signaux acoustiques peut être réglé. Pour cela, presser le bouton de volume bleu lorsque la lampe est éteinte et réduire ou couper le volume à l‘aide du bouton gauche de sélection du programme. Pour ­réactiver les signaux acoustiques ou augmenter le ­volume, presser le bouton de volume bleu, puis le bouton droit de sélection du programme. 26 Intensité lumineuse L‘intensité lumineuse est maintenue à un niveau ­cons-tant pendant l‘utilisation. Si le conducteur de lumière 10mm fourni avec la lampe est utilisé, l‘intensité lumineuse est d‘environ 1200 mW/cm2 ± 10%. L‘utilisation d‘embouts lumineux autres que ceux fournis a une influence directe sur l‘intensité lumineuse émise. Dans les embouts lumineux standards aux parois ­parallèles (10 mm), le dia-mètre est égal aux deux extrémités. Dans les embouts lumineux turbo rétrécis à leur extrémité (10>8 mm, 6>2 mm Pin-Point), le dia­ mètre de la partie arrière est plus large que celui de la ­fenêtre d‘émission de lumière. La lumière bleue incidente est ainsi concentrée sur une plus petite surface, ce qui augmente l‘intensité lumineuse émise. Les embouts lumineux Pin-Point conviennent pour la polymérisation de petites surfaces, comme par ex. la ­fixation de facettes avant l‘élimination des excès. Pour des polymérisations de zones plus larges, il est n­ écessaire de changer l‘embout lumineux. Mesurer l‘intensité lumineuse L‘intensité lumineuse de la Bluephase N et son conducteur de lumière 10 mm peuvent être contrôlés à l‘aide du radiomètre Bluephase Meter. Si la valeur mesurée ne correspond pas à l‘intensité ­lumineuse attendue, procéder de la façon suivante : – Contrôler le programme de polymérisation sélectionné – Nettoyer le radiomètre avec un chiffon humide – Retirer le conducteur de lumière et nettoyer la fenêtre d‘émission de lumière de la pièce à main à l‘aide d‘un coton imbibé d‘alcool. – Nettoyer si possible les conducteurs de lumières souillés (voir “Maintenance et nettoyage”) – Remplacer tout conducteur de lumière endommagé par un nouveau Si ces mesures n‘améliorent pas le résultat, veuillez ­contacter votre distributeur ou le service après-vente Ivoclar Vivadent. Maintenance et nettoyage Pour des questions d’hygiène, il est recommandé d’utiliser une gaine de protection jetable pour chaque patient. Les surfaces contaminées du dispositif ainsi que les embouts de protection oculaire doivent être désinfectés (par exemple FD333, FD366/Dürr Dental, Incidin Liquid/ Ecolab) et ­l’embout lumineux stérilisé avant chaque ­utilisation si le dispositif est utilisé sans une gaine de ­protection jetable. Assurez-vous qu‘aucun liquide ni autre substance ne pénètre dans la pièce à main ou la base de chargement pendant le nettoyage (risque d‘électrocution). Déconnecter la base de chargement du courant lors des opérations de nettoyage. Entourage plastique Essuyer la pièce à main avec une solution sans aldéhyde. Ne pas utiliser de solutions corrosives (ex. solutions à base d‘essence d‘orange ou contenant plus de 40% d‘éthanol), de solvants (ex. l‘acétone), ou d‘instruments pointus, qui risqueraient d‘endommager ou d‘abîmer le plastique. Nettoyer le plastique souillé avec une solution savonneuse. Conducteur de lumière Avant de nettoyer et/ou de désinfecter le conducteur de lumière, celui-ci doit subir un traitement préalable. Ceci est valable aussi bien pour un nettoyage et une ­désinfection mécaniques que manuels. Traitement préalable – Éliminer toute contamination immédiatement après utilisation ou dans les 2 heures suivantes. Pour ce faire, rincer minutieusement le conducteur de lumière sous l‘eau courante (pendant au moins 10 secondes). Il est également possible d‘utiliser une solution de désinfection sans aldéhyde adaptée pour éviter la fixation de sang. – Pour éliminer la contamination manuellement, utiliser un pinceau ou un chiffon doux. Un composite ­partiellement polymérisé peut être éliminé avec de l‘alcool et à l‘aide d‘une spatule en plastique. Ne pas utiliser d‘objets coupants ou pointus, ceux-ci ­pour­raient endommager la surface. Nettoyage et désinfection Pour le nettoyage, immerger l’embout lumineux dans une solution de nettoyage de manière à ce qu’il soit suffisamment recouvert par le liquide (des ultrasons ou un brossage délicat avec un pinceau doux peuvent renforcer le nettoyage). Un agent de nettoyage neutre-enzymatique est recommandé. Lors du nettoyage et de la désinfection, vérifier que les agents utilisés sont exempts : – d‘acides organique, minéral et oxydant (la valeur pH minimum admissible est 5,5) – de solutions alcalines (la valeur pH maximum ­admissible est 8,5) – d‘agents oxydants (par ex. peroxydes d‘hydrogène). Enlever ensuite l’embout lumineux de la solution et le rincer minutieusement à l’eau courante (pendant au moins 10 secondes). Un nettoyage dans un stérilisateur thermique est une alternative efficace. Stérilisation Un nettoyage et une désinfection intensifs sont les ­préalables indispensables pour assurer une stérilisation efficace. Utiliser pour cela la stérilisation en autoclave. Le temps de stérilisation (temps d’exposition à température de stérilisation) est de 4 minutes à 134 °C; la pression doit être de 2 bar. Sécher l’embout lumineux stérilisé en utilisant soit le programme de séchage spécial de l’autoclave, soit de l’air chaud. L’embout lumineux a été testé pour résister jusqu’à 200 cycles de stérilisation. Vérifier ensuite que le conducteur de lumière n‘ait subi aucun dommage. Le tenir à contre-jour. Si certains ­segments deviennent noirs, c‘est que les fibres de verre sont cassées. Dans ce cas, remplacer le conducteur de lumière par un neuf. Contacts de la batterie Pour garantir une conductivité fiable et constante, éviter toute contamination des contacts de la batterie (par exemple par des résidus de composite). Pour cela, n­ ettoyer régulièrement les contacts lors de la désinfection habituelle (après chaque patient). Élimination des pièces usagées Ne pas jeter la lampe à photopolymériser avec les ordures ménagères. Le recyclage doit se faire conformément à la réglementation et aux dispositions nationales. Ne pas incinérer les batteries. 27 Que faire en cas de …? Symbole Pas d‘affichage pendant le chargement La LED de la base de chargement ne s‘allume pas 28 Causes Rectification de l‘erreur Défaut d‘électronique dans la pièce à main ou la batterie Retirer et réinsérer la batterie. Si l‘erreur persiste, remplacer la batterie par le power pack (Click & Cure). Si l‘erreur persiste, veuillez contacter votre distributeur ou le service après-vente local. L‘appareil a surchauffé ou a subi une température trop basse Laisser l‘appareil refroidir (ou le laisser remonter à température ambiante s‘il est trop froid) et essayer de nouveau après un certain délai. Si l‘erreur persiste, veuillez contacter votre distributeur ou le service après-vente local. Défaut électronique dans le batterie Retirer et réinsérer la batterie. Si l‘erreur persiste, placer la pièce à main sur la base de chargement. Si malgré cela, l‘erreur persiste toujours, remplacer la ­batterie par le power pack (Click & Cure). Veuillez contacter votre distributeur ou le service après-vente local. Batterie vide Positionner la pièce à main correctement dans la base de chargement. „Charging Battery“ s‘affiche brièvement avant le retour en mode stand-by de l‘écran. Si la batterie n‘est pas mise en charge, nettoyer les contacts. Si l‘erreur ­persiste, veuillez contacter votre distributeur ou le service après-vente local. Les contacts sont contaminés Vérifier que la pièce à main est correctement en place dans la base de chargement. Nettoyer les contacts. Information : „Charging Battery“ s‘affiche brièvement avant le retour en mode stand-by de l‘écran. Si nettoyer les contacts n‘améliore pas les résultats, retirer la batterie de la pièce à main. Séparer la batterie de la pièce à main et la mettre en charge dans la base de ­chargement. Si l‘erreur persiste, veuillez contacter votre distributeur ou le service après-vente local. Batterie en mode sécurité (circuit de protection intégré) Le Power pack n‘est pas connecté ou est défectueux. Vérifier que le power pack est correctement positionné dans la base de chargement (vérifier également la position des contacts), et qu‘il est raccordé à la prise de courant par le cordon électrique (le témoin lumineux sur le power pack s‘allume en vert s‘il fonctionne correctement). Si l‘erreur persiste, veuillez contacter votre distributeur ou le service après-vente local. Garantie / Procédure en cas de réparation La lampe Bluephase N est garantie 3 ans à partir de la date d‘achat (batterie 1 an). Des pannes dues à un matériel défectueux ou à des vices de fabrication sont réparées gratuitement pendant la durée de la garantie. La garantie ne couvre aucun ­dommage, matériel ou non, autre que ceux mentionnés. L‘appareil doit être utilisé exclusivement pour les indications définies. Tout autre usage est contre-indiqué. Le fabricant rejette toute responsabilité quant à un mauvais usage de l‘appareil. Aucune demande de garantie ne peut être acceptée dans ce genre de cas. Ceci est ­particulièrement valable en cas de : – dommage lié à un usage inadapté, spécialement dans le cas de batteries mal conservées (voir Batterie); – dommage de composant lié à l‘usure dans les ­conditions d‘utilisation normale (voir Batterie); – dommage lié à des facteurs externes, exemple un choc, une chute sur le sol; – dommage lié à des réglages ou à une installation incorrects; – dommage survenant lors du branchement de la lampe sur une source d‘énergie dont le voltage et la fréquence ne sont pas conformes à ceux inscrits sur le socle de la lampe. – dommage lié à des réparations inadaptées ou à des modifications réalisées par des réparateurs non certifiés. En cas de réclamation sous garantie, la lampe complète (pièce à main, base de chargement, cordon électrique, et le power pack) doivent être retournés en port dû au distributeur ou directement à Ivoclar Vivadent, avec la facture d‘achat. Utiliser le conditionnement d‘origine et le carton correspondant pour le transport. Les réparations doivent être réalisées uniquement par un service certifié par le service après-vente Ivoclar Vivadent. En cas de défaut ne pouvant être rectifié, veuillez contacter votre distributeur ou le service aprèsvente local (voir les adresses au dos). Une description claire du défaut ou des conditions de sa survenue ­pourront faciliter l‘identification du problème. Veuillez joindre ce descriptif quand vous retournez l‘appareil. Spécifications du produit Présentation 1 Base de chargement,1 Cordon électrique, 1 Pièce à main, 1 Batterie (Li-Po), 1 Conducteur de lumière 10 mm, 3 Embouts de protection oculaire, 1 Mode d‘emploi Accessoires Les accessoires suivants sont disponibles pour la Bluephase N : REF 656196 608538 551756 592496 659956 659681 607922 608554 Description Conducteur de lumière 10 mm, noir (N) Conducteur de lumière Pin-point 6>2 mm, noir Embouts de protection oculaire Disque de protection oculaire Batterie Bluephase N Pièce à main Bluephase N Radiomètre Bluephase Meter Housses de protection Bluephase Données techniques Voltage base de chargement 5 VDC Voltage pièce à main 3.7 VDC avec la batterie 5 VDC avec le power pack Power pack 100–240 VAC / 50-60 Hz / max. 0.4 A Fabricant: Friwo Puissance 5 VDC TYPE 15.2630 Conditions de fonctionnement Température Humidité relative Pression ambiante Dimensions de la base de chargement Poids de la base de chargement Temps de chargement Source électrique de la pièce à main Temps de batterie maximum Source lumineuse Spectre Intensité lumineuse Fonctionne Dimensions de la pièce à main Poids de la pièce à main +10 °C à +35 °C 30% à 75% 700 hPa à 1060 hPa L = 205 x W = 150 x H = 85 (mm) 250 g env. 2 h (avec une batterie vide) Batterie Li-Po env. 60 min. (avec une batterie neuve et totalement chargée) Polywave® LED 385 – 515 nm max. 1'200 mW/cm2 ± 10 % 5 min. on / 6 min. off (par intermittence) L = 260 x W = 42 x H = 120 (mm) (batterie incluse) 225 g Transport et conditions de stockage : Température -20 °C à +60 °C Humidité relative 10% à 75% Pression ambiante 500 hPa à 1060 hPa Bluephase N, doit être rangée en hauteur et dans une pièce fermée. Protéger l‘appareil des chocs importants. Batterie – Ne pas stocker à des températures supérieures à 40 °C (ou 60 °C pour une courte période). Température de stockage recommandée 15 – 30 °C. – Toujours stocker une batterie chargée – et jamais pendant plus 6 mois. 29 Estimado cliente, Una óptima polimerización es un requisito ­importante para todos los materiales fotopolimerizables con el fin de realizar duraderas restauraciones de alta calidad. La lámpara de polimerización seleccionada también juega un papel decisivo a este respecto. Por ello, nos gustaría agradecerle el que usted haya comprado Bluephase N. Bluephase N es un dispositivo médico de alta calidad que ha sido diseñado de acuerdo con los últimos estándares de la ciencia y tecnología en conformidad con importantes estándares industriales. Estas instrucciones de uso le ayudarán a poner en marcha la lámpara de forma segura, aprovechar todas sus posibilidades y asegurar una larga vida útil. Si tuviera usted alguna pregunta adicional, no dude en contactarnos (consulte las direcciones en la página del reverso). Su equipo de Ivoclar Vivadent 30 Español Índice Producto32 – Despiece – Indicadores en la base de carga – Indicadores en la pieza de mano Seguridad  – Uso previsto – Indicaciones – Aclaración de símbolos – Avisos de seguridad 34 Puesta en marcha – Base de carga – Pieza de mano – Batería – Funcionamiento con cable Click&Cure 36 Funcionamiento  – Selección del programa de polimerización y tiempo de polimerización – Función de memoria de polimerización – Inicio – Señales acústicas – Intensidad lumínica – Medición de la intensidad lumínica 39 Mantenimiento y limpieza  – Carcasa – Conductos de luz – Contactos de la batería – Desecho 41 ¿Qué sucede si …?  42 Garantía  43 Procedimiento en caso de reparación  43 Especificaciones del producto  43 31 Despiece 13 12 1 3 5 6 4 10 2 11 9 1 Pieza de mano 2 Interruptor de puesta en marcha 3 Pantalla 4 Tecla de selección de programa 5 Tecla de selección de tiempo 6 Tecla de volumen 7 Indicador de potencia 8 Base de carga 9 Cable eléctrico 10 Unidad de alimentación 11 Batería 12 Conducto de luz 10 mm 13 Cono antideslumbrante 32 7 8 Indicadores en la base de carga Indicador se ilumina azul – la base de carga está conectada a la red Indicadores en la pieza de mano Programa de polimerización y estado de funcionamiento respectivamente Indicaciones seleccionadas y tiempo de polimerización respectivamente Estado de carga de la batería 33 Seguridad Uso previsto Bluephase N es una lámpara de polimerización LED que produce luz azul de alta energía. Se utiliza para la polimerización de los materiales dentales fotopolimerizables directamente en el sillón dental. El uso previsto también incluye la observación de estas instrucciones de uso. Notas de seguridad Bluephase N es un dispositivo electrónico y un p­ roducto sanitario sujeto a las directivas IEC 60601-1 (EN 606011) y EMC, además de las Directivas de Disposi­tivos ­Médicos 93/42/EEC. El aparato cumple con las normas vigentes de la UE y está clasificado como p­ roducto LED de Clase 2. Indicaciones Con su espectro de banda ancha “multiondas”, Bluephase N está indicada para la polimerización de todos los materiales dentales fotopolimerizables en la franja de longitud de onda de 385–515 nm. Estos materiales incluyen materiales de restauración, agentes adhesivos, bases, liners, selladores de fisuras, provisionales, así como materiales de cementación para brackets y trabajos de laboratorio tales como inlays de cerámica. Aclaración de símbolos Uso no permitido Símbolos en la lámpara Doble aislamiento (El dispositivo cumple con la Clase de Seguridad II) Protección frente al choque eléctrico (aparato tipo BF) Observar las instrucciones de uso Observar las instrucciones de uso Precaución El aparato no se puede eliminar con la ­basura doméstica normal. Por favor, visite la página Web de Ivoclar Vivadent de su país, para información más detallada sobre la manera de desechar las lámparas. Reciclable 34 El aparato has sido enviado desde el fabricante en condiciones técnicas y de seguridad optimas. Para mantener dichas condiciones y asegurar un funcionamiento seguro, se deben tener en cuenta las notas y regulaciones de estas instrucciones de uso. Para evitar daños al equipo y riesgos para los pacientes, los usuarios y terceras personas, deben tener en cuenta las siguientes instrucciones de seguridad. Los materiales que se polimerizan fuera de la franja de longitud de onda de 385–515 nm (hasta la fecha no se conocen). Si usted no está seguro acerca de cierto productos, por favor pregunte al fabricante del correspondiente material. No cargar ni utilizar el aparato cerca de ­sustancias inflamables o combustibles. Los dispositivos de comunicación portátiles o móviles de alta frecuencia pueden interferir con el equipamiento médico. No está permitido el uso de teléfonos móviles durante el funcionamiento de la lámpara. Uso y Responsabilidad – Bluephase N solo se puede utilizar para el uso previsto y cualquier otro uso está contraindicado. No se puede aceptar responsabilidad por daños que resulten de un uso inadecuado o de la no observancia de las instrucciones de uso. – El usuario tiene la responsabilidad de revisar Bluephase N en cuanto a su uso e idoneidad para los usos previstos. Ello es especialmente importante si se utilizan al mismo tiempo otros equipos en inmediata proximidad de Bluephase N. – Utilice únicamente piezas de repuesto y accesorios originales de Ivoclar Vivadent (ver accesorios). El fabricante no acepta responsabilidad alguna por los daños que se puedan producir si se utilizan piezas de repuestos y accesorios de otros fabricantes. – El conducto de luz es una pieza aplicada y puede alcanzar una temperatura máxima de 45 °C (113 °F) en el punto de contacto con la pieza de mano durante su funcionamiento. Voltaje de funcionamiento Antes de la conexión, asegúrese de que: a) el voltaje indicado en la placa de características ­cumpla con el suministro energético local. b) la unidad haya alcanzado la temperatura ambiente. Si se utilizan por separado la batería y la unidad alimentadora, e.g. durante la puesta en marcha o el funcionamiento con cable Click&Cure, se debe evitar el contacto con los pacientes o terceras partes. No toque los contactos expuestos de la batería o de la clavija de conexión (unidad alimentadora). Supuesto de seguridad perjudicial Si se sospecha que el funcionamiento ya no puede ser seguro, se debe desenchufar de la red eléctrica para ­evitar un funcionamiento accidental. Esto se podría dar, por ejemplo, si el aparato está visiblemente dañado o ya no funciona correctamente. La interrupción completa de electricidad solo queda asegurada desenchufando el ­mismo de la red. Protección ocular Se debe evitar la exposición directa o indirecta de los ojos. La exposición prolongada de las lámparas es desagradable para la vista y puede producir daños. Por ello, se recomienda el uso de los conos antideslumbrantes incluidos. Los individuos que por lo general sean sensibles a la luz, los que tomen medicamentos fotosensibili- zantes o los que se hayan sometido a cirugía ocular, o las personas que trabajen con el aparato o en su proximidad durante largos períodos de tiempo no deberán exponerse a la luz de este dispositivo y llevar gafas protectoras (naranjas) que absorben la luz por debajo de los 515 nm. Batería Protección: Utilice únicamente piezas de repuestos originales, especialmente baterías de Ivoclar Vivadent y bases de carga. No cortocircuite la batería. No lo almacene a temperaturas por encima de los 40º C/104º F (o 60º C/ 140º F durante un corto período de tiempo). Siempre almacene las baterías cargadas. El período de almacenamiento no debe exceder los 6 meses. Puede explotar si se elimina en fuego. Por favor, tenga en cuenta que la bateria de polímero de litio puede reaccionar con explosión, fuego, y producir humo en caso de una manipulación inapropiada y daños mecánicos. Baterias de polímero de litio dañadas, no deben volver a utilizarse Los electrolitos y gases de electrolitos liberados durante explosión, fuego y desarrollo de humo son tóxicos y corrosivos. En caso de contacto con los ojos y piel lavar inmediatamente con abundante cantidad de agua. Evitar la inhalación de humos. Acudir a un especilaista ­inmeditamente en caso de indisposición. Desarrollo térmico Al igual que sucede con todas las lámpara de alto ­rendimiento, la alta intensidad lumínica provoca cierto desarrollo térmico. La prolongada exposición de áreas próximas a pulpa y tejidos blandos puede provocar daños irreversibles o reversibles. Por ello, se deben tener en cuenta los tiempos de polimerización recomendados y se deben evitar tiempos de polimerización ininterrumpidos de más de 20 segundos sobre la misma superficie dental, así como el contacto directo con la gingiva, membrana mucosa dental o piel. Si fuera necesario, polimerice restauraciones indirectas en intervalos intermitentes de 20 segundos cada vez o utilice refrigeración externa con pistola de aire. 35 Puesta en marcha Revise el suministro en cuanto a su integridad y cualquier posible daño de transporte (ver forma de suministro). Si faltaran componentes o estuvieran dañados, contacte inmediatamente con su distribuidor o su servicio técnico local. Pieza de mano Antes de insertar el conducto de luz, retire el folio protector en la correspondiente apertura de la pieza de mano. Base de carga Antes de conectar el dispositivo, asegúrese de que el voltaje indicado en la placa de características cumpla con el suministro energético local. La placa de características está situada en la parte inferior de la base de carga. Gire ligeramente el conducto de luz mientras se inserta en la pieza de mano. Seguidamente, coloque el cono antideslumbrante en el conducto de luz. Coloque la estación de carga encima de una mesa adecuada y plana. Retire el folio protector de la clavija de conexión de la unidad de alimentación. Conecte la clavija de conexión de la unidad de alimentación en el enchufe en la parte inferior de la base de carga. Inclínela ligeramente y aplique una ligera presión hasta que escuche como encaja correctamente. Conecte el cable eléctrico al suministro energético y a la unidad de alimentación. El indicador de potencia ON en el lado izquierdo de la carcasa se ilumina en azul (ver indicadores en la base de carga). 36 Batería Recomendamos cargar completamente la batería antes de su primer uso. Si la batería está completamente ­cargada, presenta una capacidad de polimerización de aproximadamente 60 minutos. Estado de carga de la batería Con la pieza de mano conectada, se muestra el actual estado de carga en la pantalla como sigue: Batería completamente cargada (capacidad de polimerización de aproximadamente 60 minutos) Batería medio llena Reserva (La última barra en el indicador de batería está rojo en modo de reserva. La batería se debe cargar lo antes posible.) Deslice la batería recta en la pieza de mano hasta que escuche que encaja perfectamente. Cuando la bateria se está cargando, aparece brevemente en el display „Charging Battery“ antes de que la pantalla regrese al modo stand-by Colocar la pieza de mano con suavidad en el apoyo correspondiente de la base de carga. Si se utilizó la ­funda, por favor retírela antes de cargar la bateria. Si fuera posible, utilice la lámpara siempre con una batería completamente cargada, ya que ello prolongará la vida útil. Por ello se recomienda colocar la pieza de mano en la base de carga después de cada paciente. El tiempo de carga para baterias vacias es de 2 horas. Batería vacía Si la batería está completamente vacía, la pieza de mano se desconecta automáticamente a modo stand-by. La lámpara ya no se puede conectar y el programa y tiempo de polimerización tampoco se pueden ajustar. Sin embargo, la pieza de mano se puede utilizar en modo de funcionamiento con cable Click&Cure. Ya que la batería es un consumible, ésta se debe ­re­emplazar después de su típico ciclo útil de 2½ años. Ver etiqueta en la batería para comprobar la edad de la misma. 37 Funcionamiento Funcionamiento con cable Click&Cure En cualquier momento se puede utilizar Bluephase N con funcionamiento con cable, pero especialmente si la ­batería está completamente vacía. Para dicho fin, retire la batería de la pieza de mano y seguidamente retire la unidad alimentadora de la parte inferior de la base de carga. No tire del cable eléctrico. Desinfecte o autoclave las superficies contaminadas de la lámpara de polimerización, así como los conductos de luz y conos antirreflectantes antes de cada uso. Además, asegúrese de que la intensidad lumínica estipulada permita una adecuada polimerización. Para ello, revise el conducto de luz en cuanto a contaminación o daños, así como la intensidad lumínica en intervalos regulares (e.g. con el radiómetro de Ivoclar Vivadent “Bluephase Meter”). Selección del programa y tiempo de polimerización El programa de polimerización y el tiempo de polimerización se pueden ajustar individualmente. Bluephase N está equipada con los tres siguientes programas de polimerización para las diferentes indicaciones. El programa de polimerización deseado se selecciona con las teclas de selección de programa. La pantalla cambia como ­corresponde (ver indicadores en la pieza de mano). Después de la primera puesta en marcha, se han predeterminado los siguientes ajustes: – HIGH (Potencia alta): 10 segundos – LOW (Potencia baja): 10 segundos – SOFT (Inicio suave): 15 segundos. El tiempo de polimerización deseado se elige utilizando las teclas de selección de tiempo. Los usuarios pueden elegir entre 5, 10, 15, 20 y 30 segundos. Inserte la clavija de conexión en la pieza de mano hasta que oiga y sienta que ha encajado en su posición. Durante la operación con cable, la base de carga no puede cargar la batería, ya que no está conectada a una fuente de energía. El aparato está completamente desconectado del suministro de alimentación sólo cuando el cable eléctrico ha sido desconectado del enchufe. 38 Tenga en cuenta las instrucciones de uso del material aplicado a la hora de seleccionar el tiempo de polimerización. Las recomendaciones de polimerización para los materiales de composites se aplican para todos los colores y, si no se indica lo contrario en las instrucciones de uso del material pertinente, a un grosor de capa máximo de 2 mm. Por lo general, estás recomendaciones se aplican a situaciones en las que el extremo del conducto de luz se coloca directamente sobre el material que se va a polimerizar. Si se aumenta la distancia entre la fuente de luz y el material, el tiempo de polimerización se ampliará de forma correspondiente. Por ejemplo, si la distancia al material es de 9 mm, la salida efectiva de luz se reduce en aproximadamente un 50%. En este caso, el tiempo de polimerización recomendado se debe duplicar. Programas de polimerización Programa HIGH POWER Consistente alta intensidad lumínica para la polimerización de materiales de restauración y cementación para restauraciones directas e indirectas. 1.200 mW/cm 2 0 t [s] 650 mW/cm2 0 Programa LOW POWER Reducida intensidad lumínica con reducido desarrollo térmico para la polimerización de adhesivos, liners, y materiales de restauración en áreas próximas a pulpa al restaurar cavidades Clase V. Materiales de restauración Composite • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow Todos los composites convencionales • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Compómeros 3) Compoglass F / Compoglass Flow Restauraciones indirectas / Materiales de cementación Variolink II Base 4) / Variolink Veneer / Variolink N Base 4) / Variolink N Clear Veneer / Variolink Ultra 5) / Dual Cement 5) / Variolink II 5) Otros Helioseal / Helioseal F / Helioseal Clear Monopaque MultiCore Flow / Multicore HB Heliosit Orthodontic Telio Add-On Flow Telio Stains IPS Empress Direct Color IPS Empress Direct Opaque Tiempo de polimerización Adhesivos AdheSE / AdheSE One F ExciTE / ExciTE DSC Tetric N-Bond / Tetric N-Bond Self-Etch Heliobond (Syntac) Materiales provisionales Telio CS Inlay/Onlay Systemp.inlay/onlay Fermit / Fermit N Telio CS Link / Systemp.link Otros Heliosit Orthodontic Vivaglass Liner Tiempo de polimerización Materiales de restauración Composite • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow Todos los composites convencionales • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Compómeros 3) Compoglass F / Compoglass Flow Tiempo de polimerización 10 seg. 15 seg. 10 seg. 20 seg. por mm cerámica: 10 seg. por segmento 10 seg. 20 seg. 20 seg. 10 seg. 15 seg. 10 seg. 10 seg. 20 seg. 10 seg. 10 seg. 10 seg. 10 seg. 20 seg. por segmento 20 seg. 20 seg. t [s] Programa SOFT START Aumento gradual de la intensidad lumínica con reducida tensión de polimerización y reducido desarrollo térmico para la polimerización de materiales de restauración. 1.200 mW/cm2 650 mW/cm2 0 Intensidad de luz 1.200 mW/cm2 15 seg. Tiempo de exposición para composites Tiempo de exposición para Tetric EvoCeram / IPS Empress Direct 10 seg. 5 15 seg. 20 seg. ­ 15 seg. 20 seg. t [s] 1) Aplicar hasta un máximo de 2 mm si en las instrucciones de uso no recomiendan ninguna otra acción (como es el caso, ej. de colores de dentina) 2) Aplicar hasta un máximo de 4 mm si en las instrucciones de uso no recomiendan ninguna otra acción (como es el caso, ej. de colores de dentina) 3) Aplicar a un grosor de capa máximo de 3mm 4) Aplicar fotopolimerización (solo en pasta base) 5) Aplicar a polimerización dual 39 Función de memoria de polimerización Los últimos ajustes utilizados, junto con la combinación del programa y tiempo de polimerización, se almacenan automáticamente. Inicio La lámpara se conecta con la tecla “Start”. Una vez transcurrido el tiempo de polimerización seleccionado, el programa de polimerización finaliza automáticamente. Si se desea, la lámpara se puede desconectar antes de finalizar el tiempo de polimerización ajustado, volviendo a pulsar la tecla “Start”. El ventilador se activa al mismo tiempo que la lámpara. Una vez transcurrido el tiempo de polimerización, el ventilador sigue funcionando durante cierto tiempo para refrigerar el aparato. La batería no se debe retirar mientras el ventilador siga funcionando. Señales acústicas Las señales acústicas son audibles para las siguientes funciones: – Inicio (Parada) – Cada 10 segundos – Cambio de programa – Cambio de tiempo de polimerización – Conexión de la batería a la base de carga – Inserción de la batería – Mensaje de error Si se desea, se puede ajustar el volumen de las señales acústicas. Para ello, pulse la tecla de volumen azul mientras que la lámpara está apagada y reduzca el volumen o desconecte las señales acústicas con la tecla de programa izquierda. Si desea conectar de nuevo las señales acústicas o incrementar el volumen, pulse la tecla azul de volumen y entonces, pulse la tecla derecha de selección de programa. 40 Intensidad lumínica La intensidad lumínica se mantiene a un nivel consistente durante el funcionamiento. Si se utiliza el conducto de luz adjunto de 10 mm, la intensidad lumínica debe ­calibrarse a 1200 mW/cm2 ± 10%. La utilización de otros conductos de luz diferentes de los suministrados tiene una influencia directa sobre la intensidad lumínica emitida. En conductos de luz de paredes paralelas (10 mm), el diámetro es igual en ambos extremos. En conductos de luz que focalizan (conducto de luz 10>8 mm, conducto Pin Point 6>2 mm), el diámetro del extremo interior es mayor que el de la ventana de emisión de luz. Así, el haz de luz azul incidente se reduce a una superficie más pequeña, lo que aumenta la intensidad lumínica emitida. Los conductos de luz Pin-Point están indicados para la polimerización de zonas reducidas, tales como la fijación de carillas antes de la eliminación del sobrante de material. Para una polimerización correcta completa, es necesario cambiar el conducto de luz. Medición de la intensidad lumínica La intensidad lumínica de Bluephase N y el conducto de luz adjunto de 10 mm se pueden revisar con Bluephase Meter. Si el valor medido no se corresponde a la intensidad lumínica esperada, siga los siguientes pasos: – Revise el programa de polimerización seleccionado – Limpie la posible suciedad del fotosensor (radiómetro) – Retire el conducto de luz y limpie la boquilla de emisión de luz de la pieza de mano con una torunda de algodón humedecida en alcohol. – Limpie la posible suciedad del conducto de luz (consultar mantenimiento y limpieza) – Reemplace el conducto de luz dañado con uno nuevo. Si estas medidas no mejoran los resultados, por favor póngase en contacto con su distribuidor o con su servicio técnico local. Mantenimiento y limpieza Por razones de higiene, se recomienda utilizar un protector plástico desechable para cada paciente. Asegúrese de colocar el protector bien ajustado al conducto de luz. Si no utiliza protectores plásticos desechables, des­infecte las superficies contaminadas del dispositivo y los conos antideslumbrantes (FD 366/Dürr Dental, Incidin L­ iquid/ Ecolab) y esterilice el conducto de luz antes de cada uso. Durante la limpieza del dispositivo, asegúrese de que ningún líquido ni sustancia externa penetra en la pieza de mano, la base de carga y especialmente al transformador (riesgo de descarga eléctrica). Desconecte la base de carga de la fuente de alimentación para limpiarla. Cubierta Limpie la pieza de mano y el soporte de la pieza de mano con una solución desinfectante común, libre de aldehídos. No limpie con soluciones desinfectantes a­ lta­mente agresivas (por ejemplo, soluciones a base de aceite de naranja o con un contenido de etanol superior al 40 %), solventes (por ejemplo, acetona) o instrumentos punzantes que podrían dañar o rayar el plástico. Limpie las piezas plásticas con una solución jabonosa. Conducto de luz Antes de limpiar y/o desinfectar el conducto de luz, es ne­cesario realizar un tratamiento previo. Esto se aplica tanto a la limpieza como a la desinfección, ya sea automática o manual. Tratamiento previo – Quite la suciedad visible inmediatamente después del uso o dentro de las 2 horas posteriores. Para ello, ­enjuague abundantemente el conducto de luz con agua corriente (durante, por lo menos, 10 segundos). También puede utilizar una solución desinfectante libre de aldehídos para evitar que se adhiera la sangre. – Para eliminar la suciedad en forma manual, utilice un cepillo de cerdas blandas o un paño suave. Los com­ puestos parcialmente polimerizados pueden eliminarse con alcohol y una espátula de plástico, si es necesario. No utilice objetos afilados o puntiagudos, ya que podrían rayar la superficie. Limpieza y desinfección Para limpiar, sumerja el conducto de luz en una solución de limpieza y asegúrese de cubrirlo por completo con el líquido (el ultrasonido o un ligero cepillado con un cepillo de cerdas blandas podría mejorar el efecto). Se recomienda utilizar un agente de limpieza enzimático neutro. Al limpiar y desinfectar, asegúrese de que los agentes utilizados no contengan: – ácidos orgánicos, minerales y oxidantes (el valor de pH mínimo admitido es de 5,5) – solución alcalina (el valor de pH máximo admitido es de 8,5) – agente oxidante (por ejemplo, peróxido de hidrógeno) Luego, retire el conducto de luz de la solución y e­ njuáguelo con abundante agua corriente (durante por lo menos 10 segundos). La limpieza con un desinfectante ­térmico es una alternativa eficaz. Esterilización Durante la fase de limpieza y desinfección, es impres­ cindible asegurarse de que la posterior esterilización sea eficaz. Para ello, utilice únicamente el método de esterilización por autoclave. El tiempo de esterilización (tiempo de exposición a temperatura de esterilizado) es de 4 minutos, a 134 °C (273 °F); la presión debe ser de 2 bares (29 psi). Seque el conducto de luz esterilizado con el programa de secado especial del autoclave a vapor o aire caliente. El conducto de luz ha sido probado para hasta 200 ciclos de esterilización. Luego, controle que el conducto de luz no esté dañado. Sujételo a contraluz. Si los segmentos individuales se ven negros, las fibras de vidrio están quebradas. En ese caso, cambie el conducto de luz por uno nuevo. Contactos de la batería Para asegurar una conductividad confiable en todo momento, mantenga los contactos de la batería libres de posibles agentes contaminantes (por ejemplo, residuos de composite). Para ello, limpie los contactos utilizados regularmente durante la desinfección de limpieza normal (después de cada paciente). Eliminación La lámpara de polimerización no debe desecharse junto con los residuos urbanos. Elimine las baterías fuera de uso y las lámparas de poli­merización de acuerdo con las normas legales correspondientes de su país. Las ­baterías no deben incinerarse. 41 ¿Qué sucede si …? Símbolo No Display durante la carga Causas Defecto electrónico en la pieza de mano o batería Retire y vuelva a colocar la batería. Si el error persiste, ­reemplace la batería con la unidad alimentadora (Click & Cure). Si el error persiste, por favor contacte con su distribuidor o su servicio técnico local. El aparato está sobrecalentado o sobre enfriado Deje que el aparato se refrigere (o que alcance temperatura ambiente si estuviera sobre enfriado) y vuelva a intentarlo después de un tiempo. Si el error persiste, por favor contacte con su distribuidor o servicio técnico local. Defecto electrónico en la batería Retire y vuelva a colocar la batería. Si el error persiste, coloque el aparato en la base de carga. Si a pesar de ello, el error persiste, reemplace la batería con la unidad alimentadora (Click & Cure). Por favor, contacte con su proveedor o servicio técnico local. Batería vacía Posición de la pieza de mano en la base de carga. Aparece brevemente en el display „Charging Battery“ antes de que la pantalla regrese al modo stand-by. Si la batería no se carga, se deben limpiar los contactos. Si el error persiste, por favor contacte con su distribuidor o servicio técnico local. Los contactos de carga están contaminados Revise si la pieza de mano está correctamente colocada en la base de carga. Limpie los contactos de la batería. Información: Aparece „Charging Battery“ brevemente en el display antes que la pantalla regrese al modo stand-by. Bateria en modo seguridad (circuito de protección integrado) No se ilumina el LED de la base de carga 42 Subsanar el error La unidad alimentadora no está conectada o está defectuosa Si al limpiar los contactos de la batería no mejoran los resultados, retire la batería del aparato. Cargue la batería independientemente de la pieza de mano en la base de carga durante aproximadamente 10 minutos. Si el error persiste, por favor contacte con su distribuidor o su servicio técnico local. Revise si el cargador está correctamente colocado en la base de carga (revise también los contactos de carga) o si el cargador está conectado con el cable a la corriente eléctrica (si funciona correctamente, los indicadores del cargador están en verdes). Si el error persiste, por favor contacte con su distribuidor o su servicio técnico local. Garantía / Procedimiento en caso de reparación Especificaciones del producto Forma de suministro El período de garantía de Bluephase N es de 3 años ­desde la fecha de compra (1 año para la batería). Las averías que se produzcan por material defectuoso o fallos de fabricación se reparan gratis durante el período de garantía. La garantía no otorga el derecho a recuperar cualquier daño de material o no material distinto de los mencionados. El aparato solo se debe utilizar para los fines previstos. Cualquier otro uso está contraindicado. El fabricante no acepta responsabilidad derivada de un uso inadecuado y, en tales casos no se puede aceptar las reclamaciones de la garantía. Ello es particularmente válido para: – daños producidos por una manipulación inadecuada, especialmente de baterías almacenadas incorrectamente (ver datos técnicos: Condiciones de transporte y almacenamiento). – daños de los componentes resultado de utilización por debajo de los estandares de las condiciones de funcionamiento (ej baterias) – daños producidos por influencias externas, e.g. explosiones, caídas al suelo – daños producidos de una incorrecta puesta en marcha o instalación – daños producidos por conectar el dispositivo a una red eléctrica cuyo voltaje y frecuencia no cumplan con las indicadas en la placa de características – daños producidos por reparaciones inadecuadas o modificaciones que no hayan sido realizadas por servicios técnicos autorizados En caso de reclamación bajo garantía, el aparato ­completo (pieza de mano, base de carga, cordón ­eléctrico y unidad alimentadora) se debe enviar portes pagados a su distribuidor o directamente a Ivoclar ­Vivadent, junto con el justificante de compra. Utilice el envase original con los correspondientes cartones de embalaje para su transporte. Los trabajos de reparación sólo se pueden realizar en un servicio técnico autorizado por Ivoclar Vivadent. En el caso de una avería que no se pueda corregir, por favor contacte con su distribuidor o servicio técnico local (consultar direcciones en el reverso). Una clara descripción de la avería o las condiciones bajo las cuales se ha producido la misma, facilitará el diagnóstico del problema. Por favor, incluya dicha descripción cuando envíe el aparato. 1 Base de carga, 1 Cable eléctrico, 1 Unidad de alimentación, 1 Pieza de mano, 1 Batería (Li-Po), 1 Conducto de luz de 10 mm, 3 conos antideslumbrantes, 1 juego de instrucciones de uso Accesorios Se dispone de los siguientes accesorios para Bluephase N REF 656196 608538 551756 592496 659956 659681 607922 608554 Descripción Conducto de luz 10 mm, negro (N) Conducto de luz Pin Point 6>2 mm, negro Conos antideslumbrantes Pantalla antideslumbrante Bluephase N batería Bluephase N pieza de mano Bluephase Meter Bluephase fundas Datos técnicos Voltaje de funcionamiento base de carga  5 VDC Voltaje de funcionamiento pieza de mano  3.7 VDC con batería  5 VDC con unidad alimentadora Unidad alimentadora  100-240 VAC / 50-60 Hz / máx. 0.4 A  Fabricante Friwo Salida  5 VDC TIPO 15.2630 Condiciones de funcionamiento Temperatura + 10º C a 35º C (+50 a + 95º F) Humedad relativa 30% a 75% Presión ambiente 700 hPa a 1060 hPa Dimensiones de la base de carga Longitud = 205 x Ancho = 150 x Altura = 85 (mm) Peso de la base de carga 250 g Tiempo de carga apróx. 2 horas (con la batería vacía) Suministro eléctrico de la pieza de mano batería Li-Po Máximo tiempo de batería apróx. 60 min. (con una batería ueva completamente cargada). Fuente de luz Polywave® LED Franja de longitud de onda 385–515 nm Intensidad lumínica máx. 1.200 mW/cm2 ± 10% Funcionamiento 5 min. en funcionamiento / 6 minutos desconectado (intermitentemente) Dimensiones de la pieza de mano Longitud = 260 x Ancho = 42 x Alto = 120 (mm) Peso de la pieza de mano (batería incl.) 225 g Condiciones de transporte y almacenamiento: Temperatura -20º C a +60º C (-4 a 158º F) Humedad relativa 10% a 75% Presión ambiente 500 hPa a 1060 hPa Bluephase N se debe almacenar un estancias cerradas y techadas. Proteja la lámpara de golpes fuertes. Batería – No almacenar a temperaturas por encima de 40º C / 104º F (o 60º C /140º F durante cortos períodos de tiempo). Temperatura de almacenamiento recomendada 15 – 30 °C (59 – 86 °F). – Almacene siempre la batería cargada – y nunca más de 6 meses 43 Caro cliente, um endurecimento perfeito constitui a base para uma restauração de elevada qualidade e de longa duração de todos os materiais fotopolimerizáveis. Para esse facto contribui, decisivamente, o ­aparelho de polimerização escolhido. Por isso, ­constatamos com satisfação que escolheu o ­Bluephase N. Trata-se de um produto médico de elevada qualidade que é fabricado com base nas normas vigentes e segundo os conhecimentos ­técnicos e científicos actuais. As instruções de Uso explicam como se coloca o aparelho em funcionamento com segurança, como se podem aproveitar facilmente todas as suas ­possibilidades e como se faz uma boa conservação para poder usá-lo durante muito tempo. Estamos à sua disposição para qualquer questão que nos queira colocar (endereços na contracapa) A sua equipa Ivoclar Vivadent 44 Vista geral de produtos  – Lista de peças – Indicações da estação de carregamento – Indicações da peça de mão  Português Índice 46 Segurança48 – Utilização prevista – Indicação – Símbolos – Recomendações relativas à segurança   Colocação em funcionamento  50 – Estação de carregamento – Peça de mão – Bateria – Modo de cabo Click & Cure   Operação53 – Selecção do programa e do tempo de exposição – Função de memória Cure Memory – Iniciar  – Sinais acústicos – Intensidade da luz – Medição da intensidade da luz  Manutenção e limpeza  – Corpo – Condutor de luz – Contactos da bateria – Eliminação  55 E se …?  56 Garantia / Procedimento em caso de necessidade de reparação  57 Especificação do produto 57 45 Lista de peças 13 12 1 3 5 6 4 10 2 11 9 1 Peça de mão 2 Botão de ligar 3 Visor 4 Tecla de selecção do programa 5 Tecla de selecção do tempo 6 Tecla de intensidade sonora 7 Indicador de alimentação eléctrica 8 Estação de carregamento 9 Cabo de ligação à rede 10 Unidade de alimentação 11 Bateria 12 Condutor de luz 10 mm 13 Cone anti-encandeamento 46 7 8 Indicações da estação de carregamento IIndicador mostra uma luz azul - A estação de carregamento está ligada à alimentação eléctrica Indicações da peça de mão Programa de exposição ou estado de serviço Indicação seleccionada ou duração da exposição Estado de carregamento da bateria 47 Segurança Utilização prevista O Bluephase N é um aparelho de polimerização LED para a produção de luz azul de alta energia, destinando-se à polimerização de materiais dentários polimerizáveis directamente na unidade de tratamento dentária. A ­utilização prevista deste aparelho inclui também a observação das notas e regras que constam destas instruções de Uso. Recomendações relativas à segurança O Bluephase N é um dispositivo eletrônico e um produto médico que é sujeito a IEC 60601-1 (EN 60601-1) e às diretivas da EMC, bem como a Diretiva de D ­ ispositivo Médico 93/42/EEC. O aparelho está em c­ onformidade com as normas pertinentes da UE e é c­ lassificado como um produto LED Classe 2. Indicação Com o seu espectro de banda ampla “Polywave®”, o ­Bluephase N é adequado para a polimerização de todos os materiais dentários polimerizáveis na faixa de comprimentos de onda de 385 a 515 nm. Destes fazem parte materiais de obturação, bondings/adesivos, forramentos da base, forramentos, selantes de fissuras, provisórios bem como materiais de fixação para brackets e peças técnico-dentárias como, por exemplo, inlays cerâmicos. Símbolos Contra-indicação Símbolos da luz de polimerização Isolamento duplo (de acordo com Classe II de segurança) Proteção contra choque elétrico (aparelho tipo BF) Observe as Instruções de Uso A lâmpada de cura foi transportada sob condição tecnicamente segura. Para manter esta condição e assegurar uma operação segura, todas as notas e os regulamentos destas Instruções de Uso devem ser observados. Para poder evitar danos ao equipamento e riscos aos pacientes, usuários e terceiros, as seguintes instruções de segurança devem ser obedecidas. Materiais cuja polimerização é activada fora da faixa de comprimentos de onda de 385 a 515 nm (actualmente não há conhecimento deste tipo de materiais). Quando não houver certeza sobre algum produto, por favor, contacte o fabricante do respectivo material. Observe as Instruções de Uso Advertência A lâmpada de cura nunca deve ser descartada no lixo doméstico normal. A informação relativa ao descarte da luz poderá ser encontrada na homepage local da Ivoclar Vivadent. Reciclável 48 Não carregar ou usar o aparelho perto de substâncias inflamáveis ou combustíveis. Equipamentos de comunicação de alta frequência portáteis ou móveis podem interferir com o equipamento médico. Durante a ­operação não é permitido o uso de telefones móveis. Utilização e responsabilidade – O Bluephase N deve ser usado apenas para a finalidade para a qual foi projectado. Uma utilização diferente desta ou para além desta é considerada contrária à utilização prevista. O fabricante não se responsabiliza por danos resultantes de qualquer utilização que não a prevista ou de uma manipulação incorrecta. – Para além disso, o utilizador deverá, ele próprio, verificar antes da utilização do Bluephase N, se este é utilizável e adequado para os fins previstos. Isto é parti-cularmente importante quando, ao mesmo tempo, for usado na vizinhança um outro equipamento. – Só podem ser usadas peças sobresselentes e acessórios originais da Ivoclar Vivadent (ver acessórios). O fabricante não se responsabiliza por danos resultantes da utilização de outras peças ­sobresselentes ou outros acessórios. – A ponteira de luz é uma parte aplicada e pode aquecer-se a um máximo de 45°C na ­interface com a peça de mão durante a operação. Tensão de serviço Antes de ligar o aparelho, verifique se a) a tensão indicada na placa de características ­corresponde à tensão da rede de alimentação e b) o aparelho adquiriu a temperatura ambiente. Em caso de manuseamento de bateria ou fonte de ­alimentação em separado - por exemplo, durante a ­colocação em funcionamento ou o trabalho no modo de cabo Click & Cure - o contacto com o paciente ou terceiros deve ser evitado. Os contactos eléctricos expostos da bateria e da ficha de ligação (unidade de alimentação) não devem ser tocados. Suspeita de segurança afectada Caso se suspeite que uma operação segura do aparelho não é possível, desligue o aparelho da bateria ou da rede e tome as medidas necessárias para prevenir uma ­utilização acidental. Este pode ser o caso, por exemplo, quando o aparelho está visivelmente danificado ou ­quando não está a trabalhar correctamente. O aparelho só está completamente desligado da rede de alimentação quando o cabo está desligado da tomada. Protecção dos olhos Uma exposição direta ou indireta aos olhos deve ser evitada. A exposição prolongada à luz torna-se desagradável para os olhos e pode resultar em dano. Assim, o uso dos ­escudos anti-ofuscantes, que são fornecidos, é sempre recomendado. Os indivíduos que são normalmente sensíveis à luz, que tomam drogas que são foto-­ sensibilizantes, que sofreram cirurgias oculares, ou as ­pessoas que trabalham com o aparelho em sua vizinhança, durante longos períodos de tempo, não devem ficar ­expostos à luz deste dispositivo e devem usar óculos de proteção (­laranja), que absorvem luz abaixo de 515 nm. Bateria Atenção: Use apenas peças originais, nomeadamente, as baterias e as estações de carregamento devem ser da Ivoclar Vivadent. Não coloque a bateria em curto-circuito. Não guarde a bateria a temperaturas superiores a 40°C (ou durante um curto espaço de tempo a 60°C) e guarde-a sempre carregada. Não deve ser guardada mais de 6 meses. Perigo de explosão quando eliminado no fogo aberto. Favor observar que a bateria de polímero de lítio pode reagir com explosão, fogo e desenvolvimento de fumaça, nos casos de manuseio impróprio e de danos mecânicos. Baterias de polímero de lítio danificadas não deveriam mais ser usadas. Os eletrólitos e vapores de eletrólitos liberados durante a explosão, o fogo e a fumaça desenvolvida são tóxicos e corrosivos. No caso de contato com os olhos e pele, lavar imediatamente com grande quantidade de água. Evitar a inalação da fumaça. Procurar o médico, em caso de indisposição. Formação de calor Como em todas lâmpadas de potência elevada, a elevada intensidade da luz provoca calor. Uma exposição prolongada da área próxima da polpa ou do tecido mole pode provocar danos irreversíveis ou reversíveis. Por isso, os tempos de exposição devem ser cumpridos. Uma polimerização contínua superior a 20 segundos na ­mesma área e contactos directos com a gengiva, a ­mucosa oral e a pele devem ser evitados a todo o custo. As restaurações indirectas devem ser executadas com intervalos intermitentes de 20 segundos ou com uma refrigeração externa por jacto de ar. 49 Colocação em funcionamento Verifique se o aparelho está completo e não sofreu danos provocados pelo transporte (ver volume de fornecimento). Caso faltem peças ou estejam danificadas, contacte imediatamente o vendedor ou o centro de assistência da sua área. Peça de mão Antes da montagem do condutor de luz remova a folha de protecção da respectiva abertura da peça de mão. Estação de carregamento Antes de ligar o aparelho, verifique se a tensão indicada na placa de características corresponde à tensão da rede de alimentação existente. A placa de características encontra-se no lado inferior da estação de carregamento. Rode o condutor de luz ligeiramente para introduzi-lo na peça de mão. A seguir, coloque o cone anti-encandeamento no condutor de luz. Coloque a estação de carregamento numa superfície plana adequada. Remova a folha de protecção da ficha de ligação da unidade de alimentação. Introduza a ficha de ligação da unidade de alimentação no lado inferior da estação de carregamento de forma inclinada e exerça uma ligeira pressão sobre ela até ouvir e sentir que encaixou. Ligue o cabo de rede à rede de alimentação e à unidade de alimentação. O indicador de “Power ON” no lado esquerdo do corpo acende com luz azul (ver indicações da estação de carregamento). 50 Bateria Antes da primeira utilização a bateria deve ser totalmente carregada! No estado totalmente carregado, a bateria tem uma capacidade de exposição de aproximadamente 60 minutos. Estado de carregamento da bateria O respectivo estado de carregamento é indicado no visor com a peça de mão activada, da seguinte forma: Bateria totalmente carregada (capacidade de exposição de aproximadamente 60 minutos) Bateria meia carregada Introduza a bateria direita na peça de mão até ouvir e sentir que encaixou. Reserva (no modo de reserva a última barra da indicação de bateria está vermelha. A bateria deve ser carregada o mais rapidamente possível.) Quando a bateria está sendo carregada a informação: „Charging Battery“ aparece brevemente no display, antes do display retornar ao modo „stand-by“. De um modo suave, colocar a peça de mão no respectivo descanso da estação de carregamento. Se uma bucha protetora for usada, é necessário removê-la, antes de carregar a bateria. Se possível, use o aparelho sempre com a bateria totalmente carregada - assim irá prolongar a sua vida útil. Por isso recomenda-se que coloque a peça de mão na estação de carregamento após cada paciente. O tempo de carregamento para baterias vazias é de 2 horas. Bateria descarregada Quando a bateria estiver totalmente descarregada, a peça de mão passa automaticamente para o modo standby. A luz não acende, e nem o programa nem o tempo de exposição podem ser ajustados. Mas a peça de mão pode ser usada em modo de cabo Click & Cure. A bateria é uma peça de desgaste que, normalmente, tem de ser substituída passado aproximadamente 2½ anos. A idade da bateria consta no autocolante na bateria. 51 Operação Modo de cabo Click & Cure O Bluephase N pode trabalhar sempre e, nomeadamente, em caso de bateria totalmente descarregada, em modo de cabo. Para esse efeito, retire a bateria da peça de mão. A seguir desligue a unidade de alimentação, tirando a ficha de li-gação do lado inferior da estação de carregamento. Não puxe pelo cabo de rede. As superfícies contaminadas do aparelho, como os condutores da luz e os cones anti-encandeamento, devem ser desinfectadas ou autoclavadas, antes de cada utilização. Além disso, verifique se a intensidade da luz permite uma polimerização satisfatória. Inspeccione o condutor de luz regularmente, para detectar eventuais sujidades ou danificações, e verifique a intensidade da luz (por exemplo, com o radiómetro “Bluephase Meter” da Ivoclar Vivadent). Selecção do programa e do tempo de exposição O programa e o tempo de exposição podem ser ajustados individualmente. O Bluephase N dispõe de 3 programas de exposição que permitem fazer aplicações diferentes. O programa de exposição pretendido é seleccionado através dos botões de selecção do programa. O visor muda o seu aspecto (ver também indicações na peça de mão). Da primeira vez que o aparelho é colocado em serviço, ele irá apresentar a seguinte configuração, pré-ajustada na fábrica: – HIGH (High Power): 10 segundos – LOW (Low Power): 10 segundos – SOFT (Soft start): 15 segundos O tempo de exposição pretendido é seleccionado com os botões de selecção do tempo. Podem ser seleccionados 5, 10, 15, 20 e 30 segundos. Insira a ficha directamente na peça de mão até ouvir e sentir o encaixe correcto. Durante a operação a estação de carregamento não pode carregar a bateria. O aparelho só estará desligado totalmente da corrente quando o cabo estiver fora da tomada. 52 Para seleccionar o tempo de polimerização, consulte o folheto informativo do material aplicado. Nos compósitos os tempos de exposição recomendados referem-se a todas as cores e a camadas com espessuras máx. de 2 mm, salvo recomendação diferente nos ­folhetos informativos do material em questão. Estas recomendações são aplicáveis, de uma maneira geral, a exposições em que a janela de emissão da luz do condutor da luz esteja directamente sobre o material que se pretende polimerizar. Conforme aumenta a distância, os tempos de exposição têm de ser mais prolongados. Com uma distância de 9 mm a intensidade efectiva da luz fica reduzida para aproximadamente 50 %, de modo que o tempo de exposição recomendado tem de ser duplicado. Intensidade de luz 1.200 mW/cm2 Tempo de exposição para compósitos 15 seg. Tempo de exposição para Tetric EvoCeram / IPS Empress Direct 10 seg. Programas de Cura Programa HIGH POWER Intensidade de luz permanentemente elevada para a polimerização de materiais de obturação e fixação em caso de restaurações directas e indirectas 1.200 mW/cm 2 0 t [s] 650 mW/cm2 0 Programa LOW POWER Intensidade de luz reduzida com temperatura mais baixa para a polimerização de adesivos, liner e materiais de obturação na área próxima da polpa, para o tratamento de cavidades da classe V Tempo de cura 10 seg. 15 seg. 10 seg. 20 seg. por mm cerâmica: 10 seg. por segmento 10 seg. 20 seg. 20 seg. 10 seg. 15 seg. 10 seg. 10 seg. 20 seg. Adhesive AdheSE / AdheSE One F ExciTE / ExciTE DSC Tetric N-Bond / Tetric N-Bond Self-Etch Heliobond (Syntac) Materiais provisórios Telio CS Inlay/Onlay Systemp.inlay/onlay Fermit / Fermit N Telio CS Link / Systemp.link Diversos Heliosit Orthodontic Vivaglass Liner Tempo de cura Materiais de restauração Compósitos • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow Todos os compósitos convencionais • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Compômeros 3) Compoglass F / Compoglass Flow Tempo de cura 10 seg. 10 seg. 10 seg. 10 seg. 20 seg. por segmento 20 seg. 20 seg. t [s] Programa SOFT START Aumento gradual da intensidade de luz com uma tensão de contracção reduzida e temperatura mais baixa para a polimerização de materiais de obturação 1.200 mW/cm2 650 mW/cm2 0 Materiais de restauração Compósitos • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow Todos os compósitos convencionais • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Compômeros 3) Compoglass F / Compoglass Flow Restaurações indiretas / Materiais de Cimentação Variolink II Base 4) / Variolink Veneer / Variolink N Base 4) / Variolink N Clear Veneer / Variolink Ultra 5) / Dual Cement 5) / Variolink II 5) Diversos Helioseal / Helioseal F / Helioseal Clear Monopaque MultiCore Flow / Multicore HB Heliosit Orthodontic Telio Add-On Flow Telio Stains IPS Empress Direct Color IPS Empress Direct Opaque 5 15 seg. 20 seg. ­ 15 seg. 20 seg. t [s] 1) Válido para camadas com espessura máxima de 2 mm contanto que as Instruções de Uso do respectivo material não registram outra recomendação (poderá ser o caso, p.ex., com as cores de dentina). 2) Válido para camadas com espessura máxima de 4 mm contanto que as Instruções de Uso do respectivo material não registram outra recomendação (poderá ser o caso, p.ex., com as cores de dentina). 3) Válido para camadas com espessura máxima de 3 mm. 4) Válido para fotopolimerização (usar somente a pasta base). 5) Para polimerização dual. 53 Função memória e cura A última regulação usada é guardada automaticamente na combinação programa e tempo de exposição. Inicio A lâmpada é ligada com o botão Iniciar. Decorrido o tempo de exposição seleccionado, o programa de exposição t­ ermina automaticamente. Se desejar desligar a luz antes de ter decorrido o tempo fixado, pressione novamente o botão de ligar. O ventilador é activado simultaneamente com a luz. No fim do tempo de exposição, este ainda continua activo, para arrefecer o aparelho. Enquanto estiver a funcionar o ventilador, a bateria não deve ser retirada. Sinais acústicos Sinais acústicos são emitidos nas seguintes funções: – Iniciar (Parar) – De 10 em 10 segundos – Mudança de programa – Mudança do tempo de exposição – Ligação da bateria à estação de carregamento – Inserção da bateria – Mensagem de erro Se desejado, o volume de som dos sinais acústicos pode ser regulado. Para esse efeito, com a luz desligada, ­prima o botão azul do nível sonoro (ver Vista geral de produtos), e com o botão de selecção do tempo (lado esquerdo) reduza o volume de som ou desligue o som totalmente. Para ligar os sinais acústicos novamente ou para aumentar o volume de som, prima o botão azul do nível sonoro e, a seguir, o botão de selecção do tempo (lado direito). Intensidade da luz A intensidade da luz mantém-se constante durante a operação do aparelho. No condutor de luz de 10 mm ornecido, a intensidade da luz foi calibrada em 1200 mW/cm2 +/- 10%. O uso do condutor de luz diferente daquele que foi fornecido tem uma influência direta na intensidade de luz emitida. Nos condutores de luz de paredes-paralelas (10 mm), o diâmetro é igual nas duas extremidades. Nos condutores de focagem de luz (condutor de luz 10>8 mm, condutor de luz Pin-Point 6>2 mm), o diâmetro da extremidade traseira do condutor é maior que aquele da janela de emissão de luz. Deste modo, a luz azul incidente será empacotada em uma superfície menor. Isto promove o aumento da intensidade da luz que será emitida. Os condutores de luz Pin-Point são adequados para a cura em áreas confinadas, como na fixação de facetas, antes da remoção dos excessos. Para completar a poli­ merização, entretanto, será necessário trocar o condutor de luz. Medição da intensidade de luz A intensidade de luz do Bluephase N e do condutor de luz de 10 mm fornecido pode ser controlada, por exemplo, com o radiómetro “Bluephase Meter”. Se o valor medido não corresponder à intensidade de luz esperada: – Verifique o programa de exposição seleccionado – Limpe o sensor óptico do radiómetro que eventualmente pode estar sujo – Remova o condutor de luz e limpe a abertura de saída da luz na peça de mão com um cotonete embebido em álcool – Limpe o condutor de luz eventualmente sujo (ver manutenção e limpeza) – Substitua o condutor de luz eventualmente danificado por um novo Se estas medidas não produzirem o resultado desejado, contacte o seu vendedor ou o centro de assistência da sua área. 54 Manutenção e limpeza Por razões de higiene, recomendamos o uso de uma barreira de proteção descartável para cada paciente. Certifique-se de ajustar a barreira protetora perfeitamente à ponteira de luz. Desinfetar superfícies contaminadas do dispositivo e dos cones anti-dispersão (uma solução alcoólica, sem aldeídos ou fenóis por ex. FD 366/Dürr ­Dental, Incidin Liquid/Ecolab) e esterilize a ponteira de luz antes de cada uso se luvas protetoras descartáveis não forem utilizadas. ­Certifique-se de que líquidos e outras ­substâncias estranhas não entrem na peça de mão, no ­porta peça de mão e, particularmente, na fonte de energia durante a limpeza (risco de choque elétrico). Desconecte a base de recarga da tomada quando for limpála. Carcaça Limpe a peça de mão e o suporte da peça de mão com um lenço com uma solução desinfetante sem aldeído usual. Não limpe com soluções desinfetantes altamente agressivas (por exemplo, soluções à base de óleo de ­laranja ou com um teor de etanol superior a 40%), ­solventes (por exemplo, ­acetona), ou instrumentos de pontas, os quais podem danificar ou arranhar o plástico. Limpe as peças ­plásticas sujas com uma solução de sabão. Ponteira de luz Antes de limpar e/ou desinfetar a ponteira de luz, ela deverá ser prétratada. Isso se aplica tanto à limpeza e desinfecção automática quanto manual. Pré-tratamento – Remover a contaminação grosseira imediatamente após a sua utilização, ou no prazo de 2 horas. Para este fim, enxaguar bem a ponteira de luz em água ­corrente (por pelo menos 10 segundos). Como alternativa, usar uma solução desinfetante sem aldeído adequada para impedir a fixação do sangue. – Para eliminar a contaminação manualmente, use um pincel ou pano macios. Compósito parcialmente polimerizado pode ser removido com álcool e espátula plástica, se necessário. Não use objetos pontiagudos ou cortantes, pois podem riscar a superfície. Limpeza e desinfecção Para a limpeza, coloque a ponteira de luz em uma solução de limpeza e certifique-se que ela esteja suficientemente coberta com líquido (ultra-som ou escove cuidadosamente com uma escova macia possa potencializar o efeito). Um agente de limpeza enzimático neutro é recomendado. Ao limpar e desinfetar, por favor, certifique-se de que os agentes utilizados não contenham: – ácidos orgânicos, minerais e oxidantes (o valor de pH mínimo admissível é 5,5) – soluções alcalinas (o valor de pH máximo admissível é 8,5) – agentes oxidantes (por ex. peróxido de hidrogênio) Depois, remova a ponteira de luz da solução e lave bem com água corrente fresca (por pelo menos 10 segundos). Limpar em um desinfetador térmico é uma alternativa ­eficaz. Esterilização Uma limpeza e desinfecção minuciosa são imperativas para garantir que a esterilização subsequente seja eficaz. Utilize apenas esterilização em autoclave para esta finalidade. O tempo de esterilização (tempo de exposição à temperatura de esterilização) é de 4 minutos a 134 ° C, a pressão deverá ser de 2 bar. Seque a ponteira de luz esterilizada utilizando o programa de secagem especial da sua autoclave a vapor ou ar quente. A ponteira de luz foi testada até 200 ciclos de esterilização. Depois disso, verifique a ponteira de luz quanto a danos. Segure-a contra a luz. Se os segmentos individuais aparecem em preto, fibras de vidro estão partidas. Se este for o caso, substitua a ponteira de luz por uma nova. Contatos da bateria Para assegurar uma condutividade confiável em todo momento, mantenha os contatos da bateria sem qualquer contaminação possível (por ex. resíduos de compósitos). Para esta finalidade, limpe os contatos afetados regularmente durante a desinfecção usual com lenços umedecidos (após cada paciente). Descarte A lâmpada de cura não deve ser descartada como lixo urbano. Descarte as baterias e lâmpadas de polimerização inutilizáveis de acordo com os requisitos legais correspondentes em seu país. Baterias não podem ser i­ncineradas. 55 E se …? Símbolo O carregamento não é indicado no visor. O LED da estação de carregamento não acende. 56 Causa do problema Solução Defeito electrónico na peça de mão ou na bateria Remova a bateria e introduza-a novamente. Se o defeito persistir, substitua a bateria pela fonte de alimentação (Click & Cure). Se o defeito ainda continuar, contacte o seu vendedor ou o ­centro de assistência da sua área. O aparelho está demasiado quente ou demasiado frio Deixe arrefecer o aparelho (ou se este estiver demasiado frio, deixe-o adquirir temperatura ambiente), espere algum tempo e tente novamente. Se o defeito persistir, contacte o seu vendedor ou o centro de assistência da sua área. Defeito electrónico bateria Remova a bateria e introduza-a novamente. Se o aparelho ­continuar a indicar um defeito, coloque-o na estação de ­carregamento. Se, mesmo assim, o defeito persistir, substitua a bateria pela fonte de alimentação (Click & Cure). Por favor, contacte o seu vendedor ou o centro de assistência da sua área. Bateria descarregada Posicionar a peça de mão na base de carga. A informação: „Charging Battery“ aparece brevemente no display, antes do display retornar ao modo „stand-by“. Se a bateria não carregar, limpe os contactos. Se o defeito ­persistir, contacte o seu vendedor ou o centro de assistência da sua área. Os contactos de carregamento estão sujos. Verifique se a peça de mão se encontra correctamente inserida na estação de carregamento. Limpe os contactos da bateria. Informação: „Charging Battery“ aparece brevemente no display, antes do display retornar ao modo „stand-by“. Bateria em modo de segurança (circuito protetor integrado). A unidade de alimentação não está ligada ou tem algum ­defeito. Se a limpeza dos contactos da bateria não der os resultados desejados, retire a bateria do aparelho. Separada da peça de mão, carregar a bateria na estação de carregamento, durante aprox. 10 minutos. Se o defeito persistir, contacte o seu vendedor ou o centro de assistência da sua área. Verificar se a unidade de alimentação está corretamente posicionada na estação de carregamento (também confira os contatos de carregamento) ou se a unidade de alimentação está conectada ao suprimento de força através do cabo de força (display da unidade de alimentação acende a luz verde quando está ­trabalhando de modo correto). Se o defeito persistir, contacte o seu vendedor ou o centro de assistência da sua área. Garantia / Procedimento em caso de necessidade de reparação A garantia para o Bluephase N é de 3 anos (bateria 1 ano), a partir da data de compra. Em caso de avarias provocadas por defeitos de material ou fabrico, a garantia abrange a reparação gratuita do aparelho. Para além disso, a garantia não concede o direito à indemnização de eventuais danos materiais ou morais. O aparelho deve ser usado apenas para a finalidade para a qual foi projectado. Uma utilização diferente desta ou para além desta é considerada contrária à utilização prevista. Os danos daí resultantes não são abrangidos pela garantia ou pela responsabilidade civil. Nestes danos incluem-se, nomeadamente: – Danos provocados por um manuseamento incorrecto, o que se aplica, nomeadamente, a baterias não correctamente guardadas (ver Dados Técnicos: Condições de transporte e armazenagem) – Danos em peças que durante o funcionamento normal estão sujeitas ao desgaste (por exemplo, a bateria). – Danos provocados por influências exteriores, por exemplo, impacto, queda ao chão – Danos provocados por uma montagem ou instalação errada – Danos provocados pela ligação a uma tensão ou ­frequência diferente daquela que é indicada na placa de características. – Danos provocados por modificações e reparações incorrectas que não foram realizadas por uma e­ ntidade autorizada. Num caso de garantia, envie o aparelho completo (peça de mão, estação de carregamento, bateria, cabo de rede e unidade de alimentação) dentro da embalagem original, juntamente com a prova de compra, isento de custos de envio, ao revendedor ou directamente à Ivoclar Vivadent. Qualquer tipo de reparação deve ser executado, exclusivamente, por um centro de assistência Ivoclar Vivadent qualificado. Se o aparelho tiver um defeito que não pode resolver, contacte o seu vendedor ou o centro de assistência da sua área (endereços na contracapa). Uma descrição clara do defeito ou das circunstâncias que levaram ao defeito facilita a localização da avaria. Por favor, junte esta descrição ao seu aparelho. Especificação do produto Forma de apresentação 1 estação de carregamento, 1 cabo de rede, 1 unidade de alimentação, 1 peça de mão, 1 bateria (LiPo), 1 condutor de luz de 10 mm, 3 cones anti-encandeamento, 1 conjunto de instruções de Uso Acessórios Para o seu Bluephase N existem os seguintes acessórios: REF 656196 608538 551756 592496 659956 659681 607922 608554 Designação Condutor de luz 10 mm preto (N) Condutor de luz Pin-Point 6/2 mm preto (Cones) anti-encandeamento Escudo anti-encandeamento Bateria Bluephase N Peça de mão Bluephase N Radiómetro Bluephase Meter Invólucros de protecção Bluephase Dados técnicos Tensão de serviço estação de carregamento  5 VDC Tensão de serviço peça de mão  3.7 VDC com bateria  5 VDC com fonte de alimentação Unidade de alimentação  100–240 VAC / 50-60 Hz / máx. 0.4 A  Fabricante Friwo Saída  5 VDC Tipo 15.2630 Condições de operação Temperatura +10 °C a +35 °C Humidade relativa 30 % a 75 % Pressão atmosférica 700 hPa a 1060 hPa Dimensões da estação de carregamento C = 205 mm; L = 150 mm; A = 85 mm Peso da estação de carregamento 250 g Tempo de bateria aprox. 2 horas (com bateria descarregada) Alimentação eléctrica peça de mão bateria LiPo Tempo de exposição máx. aprox. 60 minutos (com bateria nova e totalmente carregada) Fonte de luz „Polywave®“ LED Faixa de comprimentos de onda 385 – 515 nm Intensidade de luz máx. 1.200 mW/cm2 ± 10 % Operação 5 min. ligado / 6 min. desligado (intermitente) Dimensões da peça de mão C = 260 mm; L = 42 mm, A = 120 mm Peso da peça de mão (inc. bateria) 225 g Condições de transporte e armazenagem: Temperatura -20 °C a +60 °C Humidade relativa 10 % a 75 % Pressão atmosférica 500 hPa a 1060 hPa Guarde o Bluephase N em ambientes fechados ou cobertos. Não exponha o aparelho a fortes vibrações. Bateria – não guardar a temperaturas superiores a 40°C (ou durante um curto espaço de tempo a 60°C), recomenda-se a conservação a 15 – 30 °C. – sempre carregada – e não mais que 6 meses 57 Say›n müflterimiz, Optimum polimerizasyon, sürekli olarak yüksek kaliteli restorasyonlar elde etmek için ›fl›kla sertleflen tüm materyallerde önemli bir gerekliliktir. Seçilen polimerizasyon ›fl›¤› da bu anlamda önemli bir role sahiptir. Bu nedenle Bluephase N cihaz›n› sat›n ald›¤›n›z için size teflekkür etmek isteriz. Bluephase N, geçerli endüstriyel standartlarla uyumlu olarak en son bilim ve teknoloji standartlar›na göre tasarlanm›fl yüksek kaliteli bir t›bbi cihazd›r. Bu çal›flt›rma talimatlar› cihaz› güvenli flekilde açman›za, performans›ndan tam olarak yararlanman›za ve cihaz› uzun süre kullanman›za yard›mc› olacakt›r. Baflka sorular›n›z varsa size memnuniyetle yard›mc› oluruz (adresler için arka sayfaya bak›n). Ivoclar Vivadent Ekibiniz 58 Ürüne genel bak›fl Parçalar›n listesi fiarj ünitesi üzerindeki göstergeler El tabancas› üzerindeki göstergeler 60 Güvenlik Amaçlanan kullan›m Kullan›m alanlar› ‹flaretler ve semboller Güvenlik notlar› 62 Cihaz›n aç›lmas› fiarj ünitesi El tabancas› Akü Click & Cure kablolu çal›flt›rma 64 Çal›flt›rma Polimerizasyon program›n›n ve polimerizasyon süresinin seçilmesi Cure Memory veri saklama fonksiyonu Bafllatma Sesli sinyaller Ifl›k fliddeti Ifl›k fliddetinin ölçülmesi 67 Bak›m ve temizleme Gövde Ifl›k problar› Akü uçlar› Cihaz›n at›lmas› 69 Bu cihaz aç›lmad›¤›nda …? 70 Garanti 71 Tamir prosedürü 71 Ürün tan›m› 71 Türkçe Türkçe ‹çindekiler 59 Parçalar›n listesi 13 12 1 3 5 6 4 10 2 11 9 1 2 3 4 5 6 7 8 9 10 11 12 13 60 El tabancas› Bafllatma dü¤mesi Ekran Program seçim tuflu Zaman seçim tuflu Ses seviyesi tuflu Güç göstergesi fiarj ünitesi Elektrik kablosu Adaptör Akü Ifl›k probu, 10 mm Yans›may› önleyici koni 7 8 fiarj ünitesi üzerindeki göstergeler Gösterge mavi renkte yan›yor Gösterge ünitesi elektrik beslemesine ba¤l› El tabancas› üzerindeki göstergeler Polimerizasyon program› ve çal›flma durumu Seçilen göstergeler ve polimerizasyon süresi Akünün flarj durumu 61 Güvenlik Amaçlanan kullan›m Bluephase N, yüksek enerjili mavi ›fl›k üreten bir LED polimerizasyon ›fl›¤›d›r. Bu cihaz, difl hekimli¤i muayenehanesinde do¤rudan kullan›lan ›fl›kla sertleflen dental materyallerin polimerizasyonunda kullan›l›r. Cihaz›n amaçlanan kullan›m› ayn› zamanda bu Kullan›m Talimatlar›nda yer alan bilgilere uyulmas›n› da kapsar. Güvenlik notlar› Bluephase N, IEC 60601-1 (EN 60601-1) ve EMC direktiflerine ve aynı zamanda 93/42/EEC sayılı Tıbbi Cihaz Direktifi'ne tabi olan bir elektronik cihaz ve tıbbi üründür. Cihaz, ilgili AB düzenlemeleri ile uyumludur ve LED Sınıf 2 ürün olarak sınıflandırılmaktadır. Kullan›m alanlar› “ploywave” genifl bantl› spektrumuyla Bluephase N, 385-515 nm dalga uzunlu¤u aral›¤›nda ›fl›kla sertleflen tüm dental materyallerin polimerizasyonu için uygundur. Bu materyaller, restorasyon materyalleri, bonding ajanlar›/adezivleri, bazlar›, linerlar›, fissür örtücüler, geçici dolgular› ve braketler için yap›flt›rma materyalleri ve porselen inlayler gibi difl laboratuvar parçalar›n› kapsar. ‹flaretler ve semboller ‹zin verilmeyen kullan›m Cihazdaki semboller ‹ki kez izole edilmifl (güvenlik s›n›f› II ile uyumlu cihaz) Elektrik çarpmas›na karfl› koruma (BF tipi cihaz) Çal›flt›rma talimatlar›na uyun Çal›flt›rma talimatlar›na uyun Dikkat Cihaz normal ev çöpüyle birlikte at›lmamal›d›r. Cihazın giderilmesiyle ilgili bilgileri kendi ulusal Ivoclar Vivadent ana sayfanızda bulabilirsiniz. Geri Dönüfltürülebilir 62 Polimerizasyon ıflık cihazı, güvenli ve teknik açıdan sa¤lam bir durumda fabrikadan çıkmıfltır. Bu durumu korumak ve risksiz bir çalıflma sa¤lamak için bu Kullanma Talimatındaki bilgilere ve kurallara uyulmalıdır. Ekipmanın zarar görmesini ve hastalar, kullanıcılar ve üçüncü taraflar açısından riskleri önlemek için afla¤ıdaki güvenlik talimatları izlenmelidir. Polimerizasyonlar›n›n 385–515 nm dalga uzunlu¤u aral›¤›n›n d›fl›nda polimerize olan materyaller (ancak bugüne kadar bu tip bir materyal bilinmemektedir). Emin olmad›¤›n›z ürünlerde, bu ürünlerin imalatç›lar›ndan bilgi alman›z tavsiye edilir. Cihazı alev alıcı veya yanıcı maddelerin yakınında flarj etmeyin ve kullanmayın. Tafl›nabilir ve mobil yüksek frekansl› iletiflim sistemleri t›bbi cihazlarda enterferansa neden olabilir. Bu nedenle cihaz çal›flt›r›ld›¤› s›rada cep telefonu kullan›lmamal›d›r. Kullan›m ve sorumluluk kapsam› – Bluephase N yaln›zca amaçland›¤› flekilde kullan›lmal›d›r. Baflka bir kullan›m ya da amac›n d›fl›nda kullan›m, kurallara uygun olarak kabul edilmemektedir. Kurallara uygun olmayan kullan›m veya amaç d›fl› kullan›m sonucunda meydana gelen zararlardan imalatç› sorumlu de¤ildir. – Kullan›c›, Bluephase N’i kullanmadan önce, cihaz›n öngörülen amaçlar için uygun oldu¤unu ve bu alanlarda kullan›labildi¤ini kendi sorumlulu¤u dahilinde kontrol etmekle yükümlüdür. Bu, özellikle cihaz›n hemen yak›n›nda ayn› anda baflka cihazlar çal›flt›r›ld›¤› zaman geçerlidir. – Yaln›zca orijinal Ivoclar Vivident yedek parçalar›n› ve aksesuarlar›n› kullan›n (bkz. Aksesuarlar bölümü). Baflka yedek parça ve aksesuarlar›n kullan›lmas›ndan do¤an zararlardan imalatç› sorumlu de¤ildir. – Iflık probu uygulamalı bir parçadır ve kullanım sırasında el aleti arayüzünde maksimum 45°C sıcaklı¤a kadar ısınabilir. Çal›flt›rma gerilimi Cihaz› açmadan önce afla¤›dakilerden emin olun: a) Tip plakas›nda belirtilen gerilim besleme flebekesindeki gerilimle ayn› olmal› ve b) cihaz ortam s›cakl›¤›na gelmelidir. Akü veya adaptör (pil paketi) ayr› olarak kullan›ld›¤›nda (örn. cihaz›n bafllat›lmas› veya Click & Cure iflleviyle kablolu çal›flt›rma s›ras›nda), cihaz›n hasta veya üçüncü kiflilerle temas etmesi önlenmelidir. Pilin aç›kta kalan kontaklar›na veya ba¤lant› fifline (adaptör) dokunmay›n. Güvenli¤in k›s›tlanmas› ihtimali Tehlikesiz bir çal›flt›rman›n art›k mümkün olmad›¤› düflünüldü¤ünde, cihaz›n fifli prizden çekilmeli, akü ile ba¤lant›s› kesilmeli ve yanl›fll›kla kullan›lmas›na karfl› emniyete al›nmal›d›r. Gözle görülür bir hasar olmas› veya do¤ru flekilde çal›flmamas› durumunda, cihaz›n çal›flt›r›lma güvenli¤i azalabilir. Elektrik flebekesinden tümüyle ay›rma ancak fiflin prizden çekilmesiyle mümkündür. Gözlerin korunmas› Gözlere dolayl› veya do¤rudan ›fl›n gelmesi önlenmelidir. Uzun süreli ›fl›n tutulmas› gözler için rahats›zl›k vericidir ve gözlerde hasarlara neden olabilir. Bu nedenle, ürünle birlikte verilen yans›ma önleyicinin kullan›lmas› tavsiye edilir. Ifl›¤a karfl› genel olarak hassas olan, ›fl›¤a karfl› hassasiyet yüzünden ilaçlar ya da fotosensitizan ilaçlar alan, göz ameliyat› geçirmifl veya uzun süredir bu cihazla veya cihaz›n yak›n›nda çal›flan kifliler cihaz›n ›fl›nlar›na maruz kalmamal› ve 515 nm dalga uzunlu¤u alt›ndaki ›fl›nlar› absorbe eden koruyucu gözlük (turuncu) takmal›d›r. Akü Dikkat: Sadece orijinal parçalar, özellikle Ivoclar Vivadent aküleri ve flarj ünitelerini kullan›n. Aküye k›sa ba¤lant› yapmay›n. Cihaz› 40°C (veya k›sa bir süre için 60°C) üzerindeki s›cakl›klarda saklamay›n. Aküler her zaman flarj edilmifl halde saklanmal›d›r. Saklama süresi 6 ay› aflmamal›d›r. Aç›k aleve at›lmas› halinde patlama tehlikesi vard›r. Lityum polimer pilin uygun olmayan muamele ve mekanik hasar durumunda patlama, yangın çıkması ve duman oluflması fleklinde reaksiyon gösterebilece¤ine dikkat edin. Hasarlı lityum polimer piller artık kullanılmamalıdır. Patlama, yangın çıkması ve duman oluflması durumunda serbest kalan elektrolitler ve elektrolit dumanları toksik ve çürütücüdür. Gözler ve ciltle temas halinde hemen bol miktarda suyla yıkayın. Dumanları solumaktan kaçının. Kendinizi iyi hissetmezseniz hemen bir doktora görünün. S›cakl›k oluflumu Tüm yüksek performansl› lambalarda oldu¤u gibi, yüksek ›fl›k fliddeti bir s›cakl›¤›n oluflmas›yla ba¤lant›l›d›r. Pulpa yak›n›ndaki alanlar›n ve yumuflak dokular›n uzun süreli maruziyetinde geri döndürülemeyen veya döndürülebilen hasar oluflabilir. Bu nedenle önerilen sertleflme sürelerine uyulmal›d›r. Ayn› difl yüzeyine 20 saniyeden uzun bir süre boyunca ›fl›k tutulmas› ve ›fl›¤›n do¤rudan difleti, a¤›z mukozas› veya deri ile do¤rudan temas etmesinden kaç›n›lmal›d›r. Gerekirse indirek restorasyonlar› 20 saniyelik aralarla polimerize edin veya hava ak›m› olan harici so¤utma kullan›n. 63 Cihaz›n aç›lmas› Teslimat s›ras›nda içeri¤in eksiksiz olup olmad›¤›n› ve olas› tafl›ma zararlar› olup olmad›¤›n› kontrol edin (bkz. teslimat kapsam›). Eksik veya zarar gören parça varsa, hemen sat›c›n›zla veya Servis Merkezinizle temasa geçin. El tabancas› Ifl›k probunu monte etmeden önce el tabancas› üzerindeki ilgili aç›kl›k üzerinde bulunan koruyucu folyoyu ç›kart›n. fiarj ünitesi Üniteyi açmadan önce tip plakas› üzerinde belirtilen gerilimin yerel elektrik beslemesi ile uyumlu oldu¤undan emin olun. Tip plakas› flarj ünitesinin alt k›sm›nda yer al›r. Ifl›k probunu el tabancas›na takt›¤›n›z s›rada hafifçe döndürün. Ard›ndan yans›ma önleyiciyi ›fl›k probuna monte edin. fiarj ünitesini uygun, düz bir masa yüzeyine yerlefltirin. Adaptörün ba¤lant› fiflindeki koruyucu folyoyu ç›kart›n. Adaptörün ba¤lant› fiflini flarj ünitesinin alt taraf›ndaki sokete tak›n. Bunun için fifli hafifçe afla¤›ya e¤in ve yerine oturdu¤unu duyana ve hissedene kadar hafif bir bas›nç uygulay›n. Elektrik kablosunu flebeke prizine ve adaptöre ba¤lay›n. Cihaz gövdesinin sol taraf›ndaki Güç ON (Aç›k) göstergesi mavi renkte yanar (bkz. fiarj ünitesi üzerindeki göstergeler bölümü). 64 Akü ‹lk kullan›m öncesinde akü tamamen flarj edilmelidir. Akü tam flarjl› haldeyken yaklafl›k 60 dakikal›k ›fl›nlama kapasitesine sahiptir. Akünün flarj durumu El tabancas› aç›k oldu¤unda mevcut flarj durumu ekranda afla¤›daki gibi gösterilir: Akü tam flarjl› (yaklafl›k 60 dakikal›k polimerizasyon kapasitesi) Akü yar› flarjl› Aküyü, klik sesiyle yerine oturdu¤unu duyana ve hissedene kadar do¤rudan el tabancas›n›n içine yerlefltirin. Yedek (Yedek modunda akü göstergesinin son çubu¤u k›rm›z› renktedir. Akü mümkün olan en k›sa sürede flarj edilmelidir.) Pil flarj edilirken, ekran bekleme moduna dönmeden önce ekranda kısa süre için „Charging Battery (Pil fiarj Ediliyor)“ görülür. Kablosuz ıflık cihazınızı flarj bazasındaki yerine yerlefltiriniz. Bir hijyen koruyucu kullan›l›yorsa, bu koruyucu akünün flarj edilmesinden önce ç›kar›lmal›d›r. Mümkün oldu¤u kadar cihaz tamamen dolu akü ile kullan›lmas› gerekir. Bu, ürünün kullan›m ömrünün uzamas›na yard›mc› olur. Bu nedenle, her hastadan sonra el tabancas›n›n flarj ünitesine yerlefltirilmesi önerilmektedir. Bofl pillerin flarj olma süresi 2 saattir. Akü bofl Akü tamamen boflald›¤›nda el tabancas› otomatik olarak bekleme moduna geçer. Bu durumda ›fl›k çal›flt›r›lamaz ve polimerizasyon program› ve süresi ayarlanamaz. Buna karfl›l›k el tabancas› Click & Cure kablolu çal›flt›rma moduyla kullan›labilir. Akü bir sarf malzeme oldu¤undan, yaklafl›k 2½ y›l olan kullan›m ömrü bittikten sonra de¤ifltirilmelidir. Akünün ömrü için akü etiketine bak›n. 65 Çal›flt›rma Click & Cure kablolu çal›flt›rma Bluephase N’i istedi¤iniz zamanda, özellikle de akünün tamamen boflald›¤› durumlarda kablolu çal›flt›rma modunda kullanabilirsiniz. Bunun için aküyü el tabancas›ndan ç›kart›n. Ard›ndan flarj ünitesinin alt k›sm›ndaki ba¤lant› fiflinden adaptörü ç›kart›n. Bu ifllem s›ras›nda elektrik kablosunu çekmeyin. Cihaz›n kontamine olmufl yüzeylerini, ›fl›k problar›n› ve yans›may› önleyici konileri her kulan›m öncesinde dezenfekte edin veya otoklavlay›n. Dahas›, ayarlanan ›fl›k fliddetinin yeterli polimerizasyona olanak sa¤lad›¤›ndan emin olun. Bunun için düzenli olarak ›fl›k probunu kontaminasyon ve hasar aç›s›ndan, ayr›ca ›fl›k fliddetini düzenli aral›klarla kontrol edin (örn. Ivoclar Vivadent radyometresi "Bluephase Meter” ile). Polimerizasyon program›n›n ve polimerizasyon süresinin seçilmesi Polimerizasyon program› ve süresi ayr› ayr› ayarlanabilir. Bluephase N, farkl› kullan›m alanlar› için 3 polimerizasyon program›na sahiptir. Program seçim tufllar›yla istedi¤iniz polimerizasyon program›n› seçebilirsiniz. Ekran buna uygun olarak de¤iflir (ayr›ca bkz. El tabancas› üzerindeki göstergeler bölümü). Cihaz ilk kez aç›ld›¤›nda flu ayarlar önceden ayarlanm›fl olacakt›r: Ba¤lant› ucunu ç›t sesi duyacak flekilde el cihaz›na yerlefltirin. – HIGH (Yüksek güç): 10 saniye – LOW (Düflük güç): 10 saniye – SOFT (Yumuflak bafllang›çl›): 15 saniye Amaçlanan polimerizasyon süresi zaman seçim tufllar›yla seçilir. Süre olarak 5, 10, 15, 20 ve 30 saniyeyi seçebilirsiniz. Kablolu kullan›m s›ras›nda flarj ünitesi, güç kayna¤›na ba¤l› olmad›¤›ndan ne pili flarj etmede ne de ›fl›k gücü ölçmede kullan›labilir. Kablolu çal›flt›rma s›ras›nda flarj ünitesi aküyü flarj edemez. Yaln›zca elektrik kablosu soketten çekildi¤i zaman cihaz›n elektrik ba¤lant›s› kesilir. 66 Polimerizasyon süresini seçerken kullan›lan materyalin kullan›m talimatlar›n› dikkate al›n. Kompozit materyaller için belirtilen polimerizasyon önerileri tüm renkler için ve, ilgili materyalin kullan›m talimatlar›nda aksi belirtilmedi¤i sürece, maksimum 2 mm’lik tabaka kal›nl›¤› için geçerlidir. Genellikle bu öneriler, ›fl›k probunun emisyon penceresinin do¤rudan polimerize edilecek materyal üzerine yerlefltirildi¤i durumlar için geçerlidir. Ifl›k kayna¤› ile materyal aras›ndaki mesafenin art›r›lmas›, polimerizasyon süresinin de buna uygun olarak uzat›lmas›n› gerektirir. Örne¤in, materyale olan mesafe 9 mm oldu¤unda, etkin ›fl›k fliddeti yaklafl›k %50 azal›r, böylece önerilen polimerizasyon süresi iki kat›na ç›kar›lmal›d›r. Polimerizasyon program› HIGH POWER Direk ve indirek restorasyonlarda kullan›lan restoratif ve simantasyon materyallerinin polimerizasyonu için sürekli yüksek ›fl›k fliddeti. 1.200 mW/cm2 0 t [s] 650 mW/cm2 0 LOW POWER V. s›n›f kaviteler restore edildi¤i s›rada pulpa yak›n›ndaki alanlarda adezivlerin, linerlar›n ve restoratif materyallerin polimerizasyonu için düflük ›s› geliflimi ile düflük ›fl›k fliddeti 1,200 mW/cm2 15 s 10 s Restoratif materyaller Kompozitler • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow Tüm geleneksel kompozitler • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Kompomerler 3) Compoglass F / Compoglass Flow ‹ndirekt restorasyonlar / simantasyon materyalleri Variolink II Base 4) / Variolink Veneer / Variolink N Base 4) / Variolink N Clear Veneer / Variolink Ultra 5) / Dual Cement 5) / Variolink II 5) Çeflitli Helioseal / Helioseal F / Helioseal Clear Monopaque MultiCore Flow / Multicore HB Heliosit Orthodontic Telio Add-On Flow Telio Stains IPS Empress Direct Color IPS Empress Direct Opaque Polimerizasyon süresi Adezivler AdheSE / AdheSE One F ExciTE / ExciTE DSC Tetric N-Bond / Tetric N-Bond Self-Etch Heliobond (Syntac) Geçici materyaller Telio CS Inlay/Onlay Systemp.inlay/onlay Fermit / Fermit N Telio CS Link / Systemp.link Çeflitli Heliosit Orthodontic Vivaglass Liner Polimerizasyon süresi Restoratif materyaller Kompozitler • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow Tüm geleneksel kompozitler • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White Kompomerler 3) Compoglass F / Compoglass Flow Polimerizasyon süresi 10 s 15 s 10 s 20 s mm seramik baflına: 10 s, yüzey baflına 10 s 20 s 20 s 10 s 15 s 10 s 10 s 20 s 10 s 10 s 10 s 10 s 20 s yüzey baflına 20 s 20 s t [s] SOFT START Restoratif materyallerin polimerizasyonu için düflük büzülme stresi ve düflük ›s› geliflimi ile ›fl›k fliddetinin kademeli olarak art›r›lmas› 1.200 mW/cm2 650 mW/cm2 0 Iflık yo¤unlu¤u Iflık yo¤unlu¤u kompozitlerin Iflık yo¤unlu¤u Tetric EvoCeram / IPS Empress Direct 5 15 s 20 s 15 s 20 s t [s] 1) ‹lgili materyalin Kullanım Talimatı herhangi baflka bir öneride bulunmuyorsa (örne¤in dentin tonları ile görülebilir) maksimum 2 mm tabaka kalınlı¤ı için geçerlidir 2) ‹lgili materyalin Kullanım Talimatı herhangi baflka bir öneride bulunmuyorsa (örne¤in dentin tonları ile görülebilir) maksimum 4 mm tabaka kalınlı¤ı için geçerlidir 3) Maksimum 3 mm tabaka kalınlı¤ı için geçerlidir 4) Iflıkla sertlefltirme için geçerlidir (sadece baz patı kullanımı) 5) Çift sertlefltirme için geçerlidir 67 Cure Memory veri saklama fonksiyonu Kullan›lan son ayarlar, polimerizasyon program› ve polimerizasyon süresiyle birlikte otomatik olarak kaydedilir. Bafllatma Bafllat tufluyla ›fl›¤› aç›n. Seçilen polimerizasyon süresi doldu¤unda, polimerizasyon program› otomatik olarak sonland›r›l›r. ‹stendi¤inde, ayarlanan süre dolmadan ›fl›k, bafllat dü¤mesine bir kez daha bas›larak kapat›labilir. Ifl›¤›n etkinlefltirilmesiyle ayn› anda fan da aç›l›r. Ifl›klama süresinin sonland›r›lmas›ndan sonra fan cihaz› so¤utmak için belirli bir süre çal›flmaya devam eder. Fan çal›flt›¤› s›rada akü ç›kar›lmamal›d›r. Sesli sinyaller Afla¤›daki fonksiyonlarda sesli sinyaller duyulabilir: – Bafllatma (Durdurma) – Her 10 saniyede bir – Program de¤iflikli¤i – Ifl›klama süresi de¤iflikli¤i – Akünün flarj ünitesine ba¤lanmas› – Akünün yerlefltirilmesi – Hata mesaj› ‹stenildi¤inde sesli sinyallerin ses seviyesi ayarlanabilir. Bunun için ›fl›k kapal› oldu¤u s›rada mavi ses seviyesi tufluna (bkz. Ürüne genel bak›fl bölümü) bas›n ve sol süre seçim tuflunu kullanarak ses seviyesini azalt›n veya sesli sinyalleri tümüyle kesin. Sesli sinyalleri tekrar açmak veya ses seviyesini art›rmak isterseniz, mavi ses seviyesi tufluna ve ard›ndan da sa¤ süre seçim tufluna bas›n. 68 Ifl›k fliddeti Ifl›k fliddeti çal›fla s›ras›nda sabit bir düzeyde tutulur. Ürünle birlikte verilen 10 mm’lik ›fl›n probu kullan›lacaksa, ›fl›k fliddeti 1200 mW/cm2 +/- %10 olarak kalibre edilmifltir. Ekte verilenin dıflında baflka bir fiber optik kullanılması durumunda, bunun çıkan ıflı¤ın yo¤unlu¤u üzerinde do¤rudan etkisi olacaktır. Paralel duvarlı fiber optiklerde (10 mm) ıflı¤ın girdi¤i pencere ile çıktı¤ı pencerenin çapı aynıdır. Fokus yapan fiber optiklerde ise (10>8 mm fiber optik, Pin-Point [i¤ne ucu] fiber optik 6>2 mm) ıflı¤ın girdi¤i pencerenin çapı çıktı¤ı pencerenin çapından daha büyüktür. Böylece, giren mavi ıflık daha küçük bir yüzeye odaklanır. Bunun sonucunda çıkan ıflının yo¤unlu¤u artar. Pin-Point (i¤ne uçlu) fiber optikler noktasal polimerizasyonlarda, örne¤in fazlalıklar alınmadan önce venerin tesbit edilmesinde, kullanılır. Komple sertlefltirme ifllemi için fiber opti¤in de¤ifltirilmesi gerekmektedir. Ifl›k fliddetinin ölçülmesi Bluephase N ile ürünle birlikte verilen 10 mm’lik ›fl›k probunun ›fl›k fliddeti Bluephase Meter kullan›larak kontrol edilebilir. Ölçülen de¤er beklenen ›fl›k fliddetiyle uyumlu de¤ilse, flu ifllemleri gerçeklefltirin: – Seçilen polimerizasyon program›n› kontrol edin. – Radyometrenin muhtemelen kirlenmifl ›fl›k sensörünü temizleyin. – Ifl›k probunu ç›kart›n ve el tabancas›ndaki ›fl›k emisyon penceresini alkol ile ›slat›lm›fl bir pamuklu çubukla temizleyin. – Muhtemelen kirlenmifl ›fl›k probunu temizleyin (bkz. Bak›m ve temizleme bölümü). – Hasar görmüfl ›fl›k probunu yenisiyle de¤ifltirin. Al›nan bu önlemler sonuçlar› iyilefltirmezse, lütfen sat›c›n›zla veya yerel Servis Merkezinizle temasa geçin. Bak›m ve temizleme Hijyenik nedenlerden dolayı her hastada tek kullanımlık bir koruyucu kılıf kullanmanızı tavsiye ederiz. Koruyucu kılıfın ıflık probuna tam olarak oturdu¤undan emin olun. Her kullanımdan önce cihazın kontamine olmufl yüzeylerini ve yansıma önleyici konileri dezenfekte edin (FD 366/Dürr Dental, Incidin Liquid/Ecolab) ve tek kullanımlık koruyucu kılıf kullanılmıyorsa, ıflık probunu sterilize edin. Temizlik sırasında el aletine, flarj ünitesine ve özellikle flarj aletine herhangi bir sıvı veya baflka yabancı bir madde girmedi¤inden emin olun (elektrik çarpması riski). Temizlerken flarj ünitesini güç kayna¤ından ayırın. Gövde El aletini ve el aleti tutucusunu aldehit içermeyen, alıflılagelmifl bir dezenfektan solüsyonla silin. Plasti¤e zarar verebilecek veya çizebilecek kuvvetli dezenfektan solüsyonlarla (örn. portakal ya¤ı esaslı veya etanol içeri¤i %40'ın üzerinde olan solüsyonlar), çözücülerle (örn. aseton) veya sivri aletlerle temizlemeyin. Kirli plastik kısımları sabunlu suyla temizleyin. Iflık probu Iflık probunu temizlemeden ve/veya dezenfekte etmeden önce ön ifllem uygulayın. Bu durum hem otomatik hem de manuel temizlik ve dezenfeksiyon için geçerlidir. Ön ifllem – Kaba kirleri kullanımdan hemen sonra veya en geç 2 saat içerisinde giderin. Bu amaçla ıflık probunu akan su altında iyice durulayın (en az 10 saniye). Alternatif olarak, kanın kurumasını önlemek için aldehit içermeyen uygun bir dezenfektan solüsyon kullanın. – Kontaminasyonu manuel olarak gidermek için yumuflak bir fırça veya yumuflak bir bez kullanın. Kısmen polimerize olmufl kompozit, alkolle ve gerekirse bir plastik spatülle çıkarılabilir. Keskin veya sivri nesneler kullanmayın çünkü bunlar yüzeyi çizebilir. Temizlik ve dezenfeksiyon Iflık probunu temizlemek için bir temizleme solüsyonuna batırın ve yeterince sıvıyla kaplan-masını sa¤layın (ultrason veya yumuflak bir fırçayla dikkatlice fırçalamak solüsyonun etkisini destekleyebilir). Nötr bir enzimatik temizleme ajanı tavsiye edilir. Temizlerken ve dezenfekte ederken, lütfen kullanılan ajanların afla¤ıdakileri içermedi¤inden emin olun: – Organik, mineral ve oksitleyici asitler (izin verilebilir minimum pH de¤eri 5,5) – Alkali solüsyon (izin verilebilir maksimum pH de¤eri 8,5) – Oksitleyici ajan (örn. hidrojen peroksit) Ardından ıflık probunu solüsyondan çıkarın ve akan su altında iyice durulayın (en az 10 sa-niye). Bir termal dezenfektörde temizlemek de etkili bir alternatiftir. Sterilizasyon Bir sonraki aflamada uygulanacak sterilizasyon iflleminin etkili olmasını sa¤lamak için iyi bir temizlik ve dezenfeksiyon flarttır. Bu amaçla sadece otoklav sterilizasyonu uygulayın. Ste-rilizasyon süresi (sterilizasyon sıcaklı¤ındaki maruziyet süresi) 134°C sıcaklıkta 4 dakikadır; basınç 2 bar olmalıdır. Sterilize edilmifl ıflık probunu buhar otoklavınızın özel kurutma programını veya sıcak hava kullanarak kurutun. Iflık probu 200 sterilizasyon döngüsüne kadar test edilmifltir. Bundan sonra, ıflık probunu hasar açısından kontrol edin. Iflı¤a do¤ru tutun. E¤er kimi bölümler siyah görünüyorsa, cam fiberler kırılmıfltır. Bu durumda, ıflık probunu yenisiyle de¤ifltirin. Pil kontakları Her zaman güvenilir iletkenlik sa¤lamak için pil kontaklarını olası kontaminasyonlardan koruyun (örn. kompozit kalıntısı). Bu amaçla, rutin olarak uygulanan silerek dezenfeksiyon ifllemi sırasında (her hastadan sonra) etkilenen kontakları düzenli olarak temizleyin. Bertaraf etme Polimerizasyon lambası evsel atık olarak bertaraf edilmemelidir. Kullanılamayacak durumdaki pilleri ve polimerizasyon lambalarını ülkenizdeki ilgili yasal gerekliliklere göre bertaraf edin. Piller yakılmamalıdır. 69 Bu cihaz aç›lmad›¤›nda ...? Sembol fiarj s›ras›nda ekranda gösterge yok. Sorunun nedeni Sorun giderme El tabancas›nda veya aküde elektronik ar›za Aküyü ç›kar›p yeniden yerlefltirin. Hata devam ederse, akünün yerine adaptörü (Click & Cure) kullan›n. Buna ra¤men hata devam ederse, lütfen sat›c›n›zla veya yerel Servis Merkezinizle temasa geçin. Cihaz afl›r› ›s›nd› veya afl›r› so¤udu Cihaz›n so¤umas›n› bekleyin (veya cihaz afl›r› so¤umuflsa oda s›cakl›¤›na gelmesini bekleyin) ve belirli bir süre sonra tekrar deneyin. Hata devam ederse, lütfen sat›c›n›zla veya yerel Servis Merkezinizle temasa geçin. Aküde elektronik ar›za Aküyü ç›kar›p yeniden yerlefltirin. Ar›za göstergesi silinmiyorsa, cihaz› flarj ünitesine tak›n. Buna ra¤men hata devam ederse, akünün yerine adaptörü (Click & Cure) kullan›n. Akü bofl El tabancasını flarj tabanında konumlandırın. Ekran bekleme moduna dönmeden önce ekranda kısa süre için „Charging Battery (Pil fiarj Ediliyor)“ görülür. Akü flarj olmuyorsa, uçlar temizlenmelidir. Hata devam ederse, lütfen sat›c›n›zla veya yerel Servis Merkezinizle temasa geçin. fiarj uçlar› kontamine oldu El tabancas›n›n flarj ünitesine do¤ru yerlefltirilip yerlefltirilmedi¤imi kontrol edin. Akü uçlar›n› temizleyin. Bilgi: Ekran bekleme moduna dönmeden önce ekranda kısa süre için „Charging Battery (Pil fiarj Ediliyor)“ görülür. Akü güvenlik modunda (entegre koruyucu devre) Adaptör ba¤l› de¤il veya ar›zal› fiarj ünitesindeki LED yanm›yor. 70 Akü uçlar›n›n temizlenmesine ra¤men hata devam ederse, aküyü cihazdan ç›kart›n. Pili ıflık cihazınızdan ayrı olarak flarj bazasında yaklaflık 10 dakika flarj ediniz. Hata devam ederse, lütfen sat›c›n›zla veya yerel Servis Merkezinizle temasa geçin. Güç ünitesinin flarj bazasına do¤ru olarak konumlandırıldı¤ını (flarj temas alanlarıda kontrol edilmelidir) veya güç ünitesinin güç kayna¤ına kablo ile ba¤lı olup olmadı¤ını (do¤ru ba¤lanmıflsa güç ünitesindeki göstergede yeflil ıflık yanacaktır) kontrol edin. Hata devam ederse, lütfen sat›c›n›zla veya yerel Servis Merkezinizle temasa geçin. Garanti / Tamir prosedürü Bluephase N garantisi sat›n alma tarihinden itibaren 3 y›ld›r (akü için 1 y›l). Materyal veya üretim hatas›ndan kaynaklanan ar›zalarda cihaz garanti süresi dahilinde ücretsiz tamir edilir. Garanti, belirtilen hasarlar d›fl›nda maddi veya maddi olmayan herhangi bir hasar›n tazmin edilmesi hakk›n› sa¤lamaz. Cihaz yaln›zca amaçland›¤› flekilde kullan›lmal›d›r. Farkl› veya amac›n› aflan kullan›ma izin verilmez. Yanl›fl kullan›mdan kaynaklanan zararlar ve bunlar›n garanti kapsam›nda tazmini için üretici herhangi bir sorumluluk kabul etmez. Bu, özellikle afla¤›dakiler için geçerlidir: – Amaca uygun olmayan kullan›mdan, özellikle do¤ru flekilde saklanmayan akülerden kaynaklanan zarar (bkz. Teknik veriler: Tafl›ma ve saklama koflullar›). – Bileflenlerde, normal çal›flt›rma s›ras›ndaki afl›nmadan kaynaklanan hasar (örn. akü). – Darbe veya yere düflürülmesi gibi d›fl etkilerden kaynaklanan hasar. – Hatal› ayar veya kurulumdan kaynaklanan hasar. – Cihaz›n, tip plakas›nda belirtilenden farkl› voltaj ve frekansa sahip bir flebekeye ba¤lanmas›ndan kaynaklanan hasar. – Yetkili olmayan servis merkezleri taraf›ndan yap›lan uygun olmayan tamirler ve modifikasyonlardan kaynaklanan hasar. Garanti durumunda cihaz›n tamam› (el tabancas›, flarj ünitesi, akü, elektrik kablosu ve adaptör) sat›fl fifliyle birlikte, orijinal ambalaj›nda ve ilgili karton malzemelerle, navlun ödenmifl olarak teslimat deposuna veya do¤rudan Ivoclar Vivadent’e gönderilmelidir. Her türlü tamir ifli yaln›zca yetkili Ivoclar Vivadent Servis Merkezleri taraf›ndan gerçeklefltirilmelidir. Gideremedi¤iniz bir ar›za durumunda lütfen sat›c›n›zla veya Servis Merkezinizle temasa geçin (adresler arka sayfadad›r). Ar›zan›n veya ar›zan›n ortaya ç›kt›¤› koflullar›n net bir dille aç›klamas› sorunun tespitini kolaylaflt›racakt›r. Lütfen cihaz› iade ederken bu aç›klamay› da ekleyin. Ürün tan›m› Teslimat kapsam› 1 fiarj ünitesi, 1 Elektrik kablosu, 1 Adaptör, 1 El tabancas›, 1 Akü (Li-Po), 1 10 mm’lik ›fl›k probu, 3 Yans›may› önleyici koni, 1 Kullan›m Talimat› Aksesuar Bluephase N cihaz›n›z için afla¤›daki aksesuarlar mevcuttur: REF Aç›klama 656196 608538 551756 592496 659956 659681 607922 608554 Ifl›k probu, 10 mm siyah (N) Nokta hedefli ›fl›k probu, 6/2 mm siyah Yans›ma önleyici (koni) Yans›may› önleyici plaka Bluephase N akü Bluephase N el tabancas› Bluephase Meter Bluephase koruyucu kolluklar Teknik veriler ‹flletme gerilimi flarj ünitesi ‹flletme gerilimi el tabancas› Adaptör Ç›k›fl T‹P 15.2630 5 VDC 3.7 VDC, akü ile 5 VDC, adaptör ile 100–240 VAC / 50-60 Hz / maks. 0.4 A Üretici: Friwo 5 VDC Çal›flma koflullar› S›cakl›k Ba¤›l Hava bas›nc› Boyutlar flarj ünitesi A¤›rl›k flarj ünitesi fiarj süresi Elektrik beslemesi el tabancas› Maks. ›fl›klama süresi Ifl›k kayna¤› Dalga boyu aral›¤› Ifl›k fliddeti ‹flletme Boyutlar El tabancas›: A¤›rl›k el tabancas› +10°C ila +35°C nem %30 ila %75 700 hPa ila 1060 hPa U = 205 mm; G = 150 mm; Y = 85 mm 250 g yakl. 2 saat (bofl aküde) Li-Po akü yakl. 60 dak. (yeni, tam flarjl› akü ile) Polywave® LED 385–515 nm maks. 1.200 mW/cm2 ± %10 5 dak. aç›k / 6 dak. kapal› (aral›kl› olarak) U = 260 mm; G = 42 mm; Y = 120 mm (akü dahil) 225 g Tafl›ma ve saklama koflullar›: S›cakl›k -20°C ila +60°C Ba¤›l nem %10 ila %75 Hava bas›nc› 500 hPa ila 1060 hPa Bluephase N kapal› veya üzeri örtülü mekanlarda saklanmal›d›r. Cihaz› güçlü sars›nt›lara maruz b›rakmay›n. Akü – 40 °C (veya k›sa bir süreli¤ine 60 °C) üzerindeki s›cakl›klarda saklamay›n. Önerilen saklama s›cakl›¤›: 15 – 30 °C. – Aküyü her zaman flarjl› halde ve en fazla 6 ay boyunca saklay›n. 71 삇ʇÂÏ˚È ÔÓÍÛÔ‡ÚÂθ, éÔÚËχθ̇fl ÔÓÎËÏÂËÁ‡ˆËfl fl‚ÎflÂÚÒfl ÓÒÌÓ‚ÓÈ ıÓÓ¯Â„Ó Í‡˜ÂÒÚ‚‡ ÂÒÚ‡‚‡ˆËË ËÁ β·˚ı Ò‚ÂÚÓÓÚ‚Âʉ‡ÂÏ˚ı χÚÂˇÎÓ‚. ëÛ˘ÂÒÚ‚ÂÌÌÛ˛ Óθ ÔË ˝ÚÓÏ Ë„‡ÂÚ Ô‡‚ËθÌ˚È ‚˚·Ó ÔÓÎËÏÂËÁ‡ˆËÓÌÌÓÈ Î‡ÏÔ˚. èÓ˝ÚÓÏÛ Ï˚ Ó˜Â̸ ‡‰˚, ˜ÚÓ Ç˚ ÓÒÚ‡ÌÓ‚ËÎË Ò‚ÓÈ ‚˚·Ó ̇ ·ÏÔ Bluephase N. ùÚ‡ ·ÏÔ‡ Ô‰ÒÚ‡‚ÎflÂÚ ÒÓ·ÓÈ ‚˚ÒÓÍÓ͇˜ÂÒÚ‚ÂÌÌ˚È ÔÓ‰ÛÍÚ Ï‰ˈËÌÒÍÓÈ ÚÂıÌËÍË, ÍÓÚÓ˚È ·˚Î ÒÍÓÌÒÚÛËÓ‚‡Ì ‚ ÒÓÓÚ‚ÂÚÒÚ‚ËË Ò ‰ÂÈÒÚ‚Û˛˘ËÏË ÌÓχÏË, ‡ Ú‡ÍÊÂ Ò Û˜ÂÚÓÏ ÒÓ‚ÂÏÂÌÌÓ„Ó ÛÓ‚Ìfl Á̇ÌËÈ Ë ÚÂıÌËÍË. Ç ËÌÒÚÛ͈ËË Í ÔË·ÓÛ Ç˚ ̇ȉÂÚ ÔÓ‰Ó·ÌÓ ‡Á˙flÒÌÂÌËÂ, Í‡Í Â„Ó ÔËÏÂÌflÚ¸, Í‡Í Ì‡Ë·ÓΠÔÓÒÚ˚Ï Ë Û‰Ó·Ì˚Ï ÒÔÓÒÓ·ÓÏ ËÒÔÓθÁÓ‚‡Ú¸ ‚ÒÂ Â„Ó ‚ÓÁÏÓÊÌÓÒÚË Ë Í‡Í Á‡ ÌËÏ Ûı‡ÊË‚‡Ú¸, ˜ÚÓ·˚ ÓÌ ÔÓÒÎÛÊËÎ Ç‡Ï Í‡Í ÏÓÊÌÓ ‰Óθ¯Â. ÖÒÎË Û Ç‡Ò ‚ÓÁÌËÍÌÛÚ ‚ÓÔÓÒ˚, Ï˚ Ò Û‰Ó‚ÓθÒÚ‚ËÂÏ Ì‡ ÌËı ÓÚ‚ÂÚËÏ (‡‰ÂÒ‡ Ç˚ ̇ȉÂÚ ̇ ÔÓÒΉÌÂÈ ÒÚ‡Ìˈ ËÌÒÚÛ͈ËË). àÒÍÂÌÌ LJ¯ Ivoclar Vivadent 72 English êyccÍËÈ êÛÒÒÍËÈ ëÓ‰ÂʇÌË é·ÁÓ ÔÓ‰ÛÍÚ‡ – èÂ˜Â̸ ÒÓÒÚ‡‚Ì˚ı ˜‡ÒÚÂÈ – èÓ͇Á‡ÌËfl Á‡fl‰ÌÓ„Ó ·ÎÓ͇ – èÓ͇Á‡ÌËfl ̇ÍÓ̘ÌË͇ 74 ÅÂÁÓÔ‡ÒÌÓÒÚ¸ – àÒÔÓθÁÓ‚‡ÌË ÔÓ Ì‡Á̇˜ÂÌ˲ – èÓ͇Á‡ÌËfl – ê‡Á˙flÒÌÂÌË Á̇ÍÓ‚ – íÂıÌË͇ ·ÂÁÓÔ‡ÒÌÓÒÚË 76 Ç‚Ó‰ ‚ ˝ÍÒÔÎÛ‡Ú‡ˆË˛ – á‡fl‰Ì˚È ·ÎÓÍ – ç‡ÍÓ̘ÌËÍ – ÄÍÍÛÏÛÎflÚÓ – ꇷÓÚ‡ ˜ÂÂÁ ͇·Âθ Click & Cure 78 àÒÔÓθÁÓ‚‡ÌË ÔË·Ó‡ – Ç˚·Ó ÔÓ„‡ÏÏ˚ ÔÓÎËÏÂËÁ‡ˆËË Ë ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË – îÛÌ͈Ëfl ÒÓı‡ÌÂÌËfl ‰‡ÌÌ˚ı Cure Memory – á‡ÔÛÒÍ – á‚ÛÍÓ‚˚ Ò˄̇Î˚ – åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl – àÁÏÂÂÌË ÏÓ˘ÌÓÒÚË Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl 81 íÂıÓ·ÒÎÛÊË‚‡ÌËÂ Ë Ó˜ËÒÚ͇ – äÓÔÛÒ – ë‚ÂÚÓ‚Ó‰ – äÓÌÚ‡ÍÚ˚ ‡ÍÍÛÏÛÎflÚÓ‡ – ìÚËÎËÁ‡ˆËfl 83 óÚÓ ‰Â·ڸ, ÂÒÎË …? 84 ɇ‡ÌÚËfl / ÑÂÈÒÚ‚Ëfl ‚ ÒÎÛ˜‡Â ÂÏÓÌÚ‡ 85 ëÔˆËÙË͇ˆËfl 85 73 èÂ˜Â̸ ÒÓÒÚ‡‚Ì˚ı ˜‡ÒÚÂÈ 13 12 1 3 5 6 4 10 2 11 9 1 ç‡ÍÓ̘ÌËÍ 2 äÌÓÔ͇ ÒÚ‡Ú‡ 3 ÑËÒÔÎÂÈ 4 ä·‚˯‡ ‚˚·Ó‡ ÔÓ„‡ÏÏ˚ 5 ä·‚˯‡ ‚˚·Ó‡ ‚ÂÏÂÌË 6 ä·‚˯‡ „ÛÎËÓ‚ÍË „ÓÏÍÓÒÚË 7 é·ÓÁ̇˜ÂÌË ÏÓ˘ÌÓÒÚË 8 á‡fl‰Ì˚È ·ÎÓÍ 9 ëÂÚ‚ÓÈ Í‡·Âθ 10 ëÂÚ‚ÓÈ ·ÎÓÍ 11 ÄÍÍÛÏÛÎflÚÓ 12 ë‚ÂÚÓ‚Ó‰ 10 ÏÏ 13 ᇢËÚÌ˚È ˝Í‡Ì 74 7 8 èÓ͇Á‡ÌËfl Á‡fl‰ÌÓ„Ó ·ÎÓ͇ ÉÓËÚ ÒËÌËÈ Ó„ÓÌÂÍ á‡fl‰Ì˚È ·ÎÓÍ ÔÓ‰Íβ˜ÂÌ Í ˝ÎÂÍÚÓÒÂÚË. èÓ͇Á‡ÌËfl ̇ÍÓ̘ÌË͇ - èÓ„‡Ïχ ÔÓÎËÏÂËÁ‡ˆËË ËÎË ÂÊËÏ ‡·ÓÚ˚ ÔË·Ó‡ - Ç˚·‡ÌÌ˚ ÔÓ͇Á‡ÌËfl ËÎË ‰ÎËÚÂθÌÓÒÚ¸ ÔÓÎËÏÂËÁ‡ˆËË - ìÓ‚Â̸ Á‡fl‰ÍË ‡ÍÍÛÏÛÎflÚÓ‡ 75 ÅÂÁÓÔ‡ÒÌÓÒÚ¸ àÒÔÓθÁÓ‚‡ÌË ÔÓ Ì‡Á̇˜ÂÌ˲ ã‡ÏÔ‡ Bluephase N – ˝ÚÓ Ò‚ÂÚÓ‰ËÓ‰Ì˚È ÔË·Ó Ò ËÁÎÛ˜ÂÌËÂÏ ÒËÌÂ„Ó ÒÔÂÍÚ‡ ‚˚ÒÓÍÓÈ ÏÓ˘ÌÓÒÚË ‰Îfl ÔÓÎËÏÂËÁ‡ˆËË Ò‚ÂÚÓÓÚ‚Âʉ‡ÂÏ˚ı ÒÚÓχÚÓÎӄ˘ÂÒÍËı χÚÂˇÎÓ‚ ÌÂÔÓÒ‰ÒÚ‚ÂÌÌÓ Û ÒÚÓχÚÓÎӄ˘ÂÒÍÓÈ ÛÒÚ‡ÌÓ‚ÍË. ä ËÒÔÓθÁÓ‚‡Ì˲ ÔÓ Ì‡Á̇˜ÂÌ˲ ÓÚÌÓÒËÚÒfl Òӷβ‰ÂÌË ÂÍÓÏẨ‡ˆËÈ Ì‡ÒÚÓfl˘ÂÈ ËÌÒÚÛ͈ËË. èÓ͇Á‡ÌËfl Å·„Ó‰‡fl ¯ËÓÍÓÔÓÎÓÒÌÓÏÛ ÔÓÎË‚ÓÎÌÓ‚ÓÏÛ ÒÔÂÍÚÛ Î‡ÏÔÓÈ Bluephase N ÏÓÊÌÓ ÔÓ‚Ó‰ËÚ¸ ÔÓÎËÏÂËÁ‡ˆË˛ ‚ÒÂı Ò‚ÂÚÓÓÚ‚Âʉ‡ÂÏ˚ı ÒÚÓχÚÓÎӄ˘ÂÒÍËı χÚÂˇÎÓ‚ ‚ ‰Ë‡Ô‡ÁÓÌ ‰ÎËÌ˚ Ò‚ÂÚÓ‚ÓÈ ‚ÓÎÌ˚ 385–515 ÌÏ. ä Ú‡ÍËÏ Ï‡ÚÂË‡Î‡Ï ÓÚÌÓÒflÚÒfl ÔÎÓÏ·ËÓ‚Ó˜Ì˚ χÚÂˇÎ˚, ·Ó̉ËÌ„Ë/‡‰„ÂÁË‚˚, ÔӉͷ‰Ó˜Ì˚ χÚÂˇÎ˚, ·ÈÌÂ˚, χÚÂˇÎ˚ ‰Îfl Á‡Ô˜‡Ú˚‚‡ÌËfl ÙËÒÒÛ, χÚÂˇÎ˚ ‰Îfl ‚ÂÏÂÌÌ˚ı ÂÒÚ‡‚‡ˆËÈ, ‡ Ú‡ÍÊ ÙËÍÒËÛ˛˘Ë ÍÓÏÔÓÁËÚ˚ ‰Îfl ·ÂÍÂÚÓ‚ Ë ÒÚÓχÚÓÎӄ˘ÂÒÍËı ÍÓÌÒÚÛ͈ËÈ, ͇Í, ̇ÔËÏÂ, ÍÂ‡Ï˘ÂÒÍË ‚Í·‰ÍË ÚËÔ‡ Inlay. ê‡Á˙flÒÌÂÌË Á̇ÍÓ‚ ç‰ÓÔÛÒÚËÏÓ ÔËÏÂÌÂÌË ëËÏ‚ÓÎ˚ ̇ ÔË·ÓÂ: Ñ‚ÓÈ̇fl ËÁÓÎflˆËfl (ÄÔÔ‡‡Ú Í·ÒÒ‡ Á‡˘ËÚ˚ II) ᇢËÚ‡ ÓÚ Û‰‡‡ ˝ÎÂÍÚÓÚÓÍÓÏ (ÚËÔ ‡ÔÔ‡‡Ú‡ BF) ëӷβ‰‡Ú¸ Ú·ӂ‡ÌËfl ËÌÒÚÛ͈ËË ëӷβ‰‡Ú¸ Ú·ӂ‡ÌËfl ËÌÒÚÛ͈ËË éÒÚÓÓÊÌÓ! èË·Ó ÌÂθÁfl ÛÚËÎËÁËÓ‚‡Ú¸ Í‡Í Ó·˚˜Ì˚È ·˚ÚÓ‚ÓÈ ÏÛÒÓ. àÌÙÓχˆË˛ ÔÓ ÛÚËÎËÁ‡ˆËË ÔË·Ó‡ ‚˚ ̇ȉÂÚ ̇ ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘ÂÈ ‰Óχ¯ÌÂÈ ÒÚ‡Ìˈ ̇ˆËÓ̇θÌÓ„Ó ‚·-Ò‡ÈÚ‡ Ivoclar Vivadent. àÒÔÓθÁÓ‚‡ÌË ÚÓθÍÓ ‚ Á‡˘Ë˘ÂÌÌÓÏ 76 ÅÂÁÓÔ‡ÒÌÓÒÚ¸ ã‡ÏÔ‡ Bluephase N – ˝ÚÓ ˝ÎÂÍÚ˘ÂÒÍËÈ ÔË·Ó Ë fl‚ÎflÂÚÒfl ωˈËÌÒÍËÏ ÔÓ‰ÛÍÚÓÏ, ÍÓÚÓ˚È ÔÓ‰ÎÂÊËÚ ÌÓÏ‡Ï IEC 60601-1 (EN 60601-1) Ë EMV, ‡ Ú‡ÍÊ ‰ÂÈÒÚ‚Û˛˘ËÏ ÌÓÏ‡Ï ÔÓ Ï‰ˈËÌÒÍËÏ ÚÓ‚‡‡Ï 93/42/EWG. èË·Ó ÒÓÓÚ‚ÂÚÒÚ‚ÛÂÚ ‰ÂÈÒÚ‚Û˛˘ËÏ ÌÓÏ‡Ï Öë Ë Í·ÒÒËÙˈËÛÂÚÒfl Í‡Í Ò‚ÂÚÓ‰ËÓ‰Ì˚È ÔË·Ó Í·ÒÒ‡ 2. èË·Ó ·˚Î ‚˚ÔÛ˘ÂÌ Á‡‚Ó‰ÓÏ ‚ ̇‰ÂÊÌÓÏ Ë ÚÂıÌ˘ÂÒÍË ·ÂÁÛÔ˜ÌÓÏ ÒÓÒÚÓflÌËË. óÚÓ·˚ ÒÓı‡ÌflÚ¸ Ë ÔÓ‰‰ÂÊË‚‡Ú¸ ˝ÚÓ ÒÓÒÚÓflÌËÂ, ‡ Ú‡ÍÊ ӷÂÒÔ˜˂‡Ú¸ ·ÂÁÓÔ‡ÒÌÛ˛ ‡·ÓÚÛ Ò ÔË·ÓÓÏ, ÒΉÛÂÚ Òӷβ‰‡Ú¸ ÂÍÓÏẨ‡ˆËË ‰‡ÌÌÓÈ ËÌÒÚÛ͈ËË. óÚÓ·˚ ËÁ·Âʇڸ Ô˘ËÌÂÌËfl ‚‰‡ Ô‡ˆËÂÌÚ‡Ï, ÔÓθÁÓ‚‡ÚÂÎflÏ Ë ÚÂÚ¸ËÏ Îˈ‡Ï, ÓÒÓ·ÂÌÌÓ ÒΉÛÂÚ Ó·‡ÚËÚ¸ ‚ÌËχÌË ̇ ÒÎÂ‰Û˛˘Ë ÔÓÎÓÊÂÌËfl: å‡ÚÂˇÎ˚, ÔÓÎËÏÂËÁ‡ˆËfl ÍÓÚÓ˚ı ‡ÍÚË‚ËÛÂÚÒfl Ì ‚ ‰Ë‡Ô‡ÁÓÌ ҂ÂÚÓ‚ÓÈ ‚ÓÎÌ˚ 385–515 ÌÏ (‚ ̇ÒÚÓfl˘ËÈ ÏÓÏÂÌÚ Ú‡ÍË χÚÂˇÎ˚ Ì ËÁ‚ÂÒÚÌ˚). Ç ÒÓÏÌËÚÂθÌ˚ı ÒÎÛ˜‡flı, ÍÓ„‰‡ Ç˚ Ì ۂÂÂÌ˚ ‚ ÔÓ‰ÛÍÚÂ, Ï˚ ÂÍÓÏẨÛÂÏ Ç‡Ï Ó·‡ÚËÚ¸Òfl Í ÔÓËÁ‚Ó‰ËÚÂβ χÚÂˇ·. èË·Ó ÌÂθÁfl ËÒÔÓθÁÓ‚‡Ú¸ Ë Áaflʇڸ ‚·ÎËÁË ‚ÓÒÔ·ÏÂÌfl˛˘ËıÒfl ËÎË ‚Á˚‚˜aÚ˚ı ‚¢ÂÒÚ‚. èÂÂÌÓÒÌ˚Â Ë ÏÓ·ËθÌ˚ ‚˚ÒÓÍÓ˜‡ÒÚÓÚÌ˚ ÍÓÏÏÛÌË͇ˆËÓÌÌ˚ Ò‰ÒÚ‚‡ ÏÓ„ÛÚ ‚ÎËflÚ¸ ̇ ‡·ÓÚÛ Ï‰ˈËÌÒÍÓÈ ÚÂıÌËÍË. èÓ˝ÚÓÏÛ Ó‰ÌÓ‚ÂÏÂÌÌÓ ËÒÔÓθÁÓ‚‡ÌË ÏÓ·ËθÌ˚ı ÚÂÎÂÙÓÌÓ‚ Ë ‡ÔÔ‡‡Ú‡ Ì ‰ÓÔÛÒÚËÏÓ. éÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚ¸ ÔÓθÁÓ‚‡ÚÂÎfl Ë ÔÓËÁ‚Ó‰ËÚÂÎfl – ã‡ÏÔ‡ Bluephase N ÏÓÊÂÚ ·˚Ú¸ ËÒÔÓθÁÓ‚‡Ì‡ ËÒÍβ˜ËÚÂθÌÓ ÚÓθÍÓ ÔÓ Ì‡Á̇˜ÂÌ˲. ã˛·Ó ‰Û„Ó ÔËÏÂÌÂÌË ‚ ËÌ˚ı ˆÂÎflı fl‚ÎflÂÚÒfl ËÒÔÓθÁÓ‚‡ÌËÂÏ Ì ÔÓ Ì‡Á̇˜ÂÌ˲. èÓËÁ‚Ó‰ËÚÂθ Ì ÌÂÒÂÚ ÓÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚË Á‡ ‚‰, Ô˘ËÌÂÌÌ˚È ‚ ÂÁÛθڇÚ ËÒÔÓθÁÓ‚‡ÌËfl ÔË·Ó‡ Ì ÔÓ Ì‡Á̇˜ÂÌ˲ ËÎË ÌÂÍÓÂÍÚÌÓ„Ó ÛıÓ‰‡ Á‡ ÌËÏ. – äÓÏ ÚÓ„Ó, ÔÓÚ·ËÚÂθ Ó·flÁ‡Ì ÔÓ‰ Ò‚Ó˛ ÓÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚ¸ ÔÓ‚ÂËÚ¸ ·ÏÔÛ Bluephase N ÔÂ‰  ËÒÔÓθÁÓ‚‡ÌËÂÏ Ì‡ ÒÓÓÚ‚ÂÚÒÚ‚ËÂ Ë ‚ÓÁÏÓÊÌÓÒÚ¸ ÔËÏÂÌÂÌËfl ‰Îfl ÔÓÒÚ‡‚ÎÂÌÌ˚ı ˆÂÎÂÈ. éÒÓ·ÂÌÌÓ ˝ÚÓ Í‡Ò‡ÂÚÒfl ÚÂı ÒÎÛ˜‡Â‚, ÍÓ„‰‡ ‚ ÌÂÔÓÒ‰ÒÚ‚ÂÌÌÓÈ ·ÎËÁÓÒÚË ÓÚ ÔË·Ó‡ Ë Ó‰ÌÓ‚ÂÏÂÌÌÓ Ò ÌËÏ ‡·ÓÚ‡˛Ú Ò ‰Û„ËÏË ‡ÔÔ‡‡Ú‡ÏË. – ÑÓÔÛÒ͇ÂÚÒfl ËÒÔÓθÁÓ‚‡Ú¸ ÚÓθÍÓ ÓË„Ë̇θÌ˚ Á‡Ô‡ÒÌ˚ ˜‡ÒÚË Ë ÔË̇‰ÎÂÊÌÓÒÚË ÙËÏ˚ Ivoclar Vivadent (ÒÏ. ‡Á‰ÂÎ èË̇‰ÎÂÊÌÓÒÚË). èÓËÁ‚Ó‰ËÚÂθ Ì ÌÂÒÂÚ ÓÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚË Á‡ ‚‰, Ô˘ËÌÂÌÌ˚È ‚ ÂÁÛθڇÚ ËÒÔÓθÁÓ‚‡ÌËfl ËÌ˚ı Á‡Ô‡ÒÌ˚ı ˜‡ÒÚÂÈ ËÎË ÔË̇‰ÎÂÊÌÓÒÚÂÈ. – ë‚ÂÚÓ‚Ó‰ – ˝ÚÓ ‡·Ó˜‡fl ˜‡ÒÚ¸ Ë ‚Ó ‚ÂÏfl ‡·ÓÚ˚ ‚ ÏÂÒÚ ÒÓ‰ËÌÂÌËfl Ò Ì‡ÍÓ̘ÌË-ÍÓÏ ÏÓÊÂÚ Ì‡„‚‡Ú¸Òfl ‰Ó χÍÒ. 45°C. ꇷӘ ̇ÔflÊÂÌË èÂ‰ ‚Íβ˜ÂÌËÂÏ Î‡ÏÔ˚ ÒΉÛÂÚ Û‰ÓÒÚÓ‚ÂËÚ¸Òfl, ˜ÚÓ a) ̇ÔflÊÂÌËÂ, Û͇Á‡ÌÌÓ ̇ Ú‡·Î˘ÍÂ, ÒÓÓÚ‚ÂÚÒÚ‚ÛÂÚ Ì‡ÔflÊÂÌ˲ ‚ ÒÂÚË Ë ·) ÔË·Ó ËÏÂÂÚ ÍÓÏ̇ÚÌÛ˛ ÚÂÏÔÂ‡ÚÛÛ. èË ‡·ÓÚ ÓÚ‰ÂθÌÓ Ò ‡ÍÍÛÏÛÎflÚÓÓÏ ËÎË ÒÂÚ‚˚Ï ·ÎÓÍÓÏ – ̇ÔËÏÂ, ÔË ‚‚Ӊ ‚ ˝ÍÒÔÎÛ‡Ú‡ˆË˛ ËÎË ÔË ‡·ÓÚ ˜ÂÂÁ ͇·Âθ Click & Cure, – ËÁ·Â„‡Ú¸ ÍÓÌÚ‡ÍÚ‡ Ò Ô‡ˆËÂÌÚÓÏ ËÎË ÚÂÚ¸ËÏË Îˈ‡ÏË. ç ‰ÓÚ‡„Ë‚‡Ú¸Òfl ‰Ó Ó„ÓÎÂÌÌ˚ı ˝ÎÂÍÚ˘ÂÒÍËı ÍÓÌÚ‡ÍÚÓ‚ ‡ÍÍÛÏÛÎflÚÓ‡ ËÎË ÒÓ‰ËÌËÚÂθÌÓ„Ó ¯ÚÂÍÂ‡ (ÒÂÚÂ‚Ó„Ó ·ÎÓ͇). åÂ˚ Ô‰ÓÒÚÓÓÊÌÓÒÚË ÖÒÎË ‚ÓÁÌËÍÎÓ Ô‰ÔÓÎÓÊÂÌËÂ, ˜ÚÓ ·ÂÁÓÔ‡ÒÌÓ ËÒÔÓθÁÓ‚‡ÌË ÔË·Ó‡ Ì‚ÓÁÏÓÊÌÓ, ÔË·Ó ÒΉÛÂÚ ÓÚÍβ˜ËÚ¸ ÓÚ ‡ÍÍÛÏÛÎflÚÓ‡ Ë ˝ÎÂÍÚÓÒÂÚË Ë ÔËÌflÚ¸ ÏÂ˚, ˜ÚÓ·˚ ÓÌ Ì ·˚Î ‚Íβ˜ÂÌ ‚ ÂÁÛθڇÚ ÌÂÓ҂‰ÓÏÎÂÌÌÓÒÚË ÔÂÒÓ̇·. í‡ÍÓ Ô‰ÔÓÎÓÊÂÌË ÏÓÊÂÚ ‚ÓÁÌËÍÌÛÚ¸, ̇ÔËÏÂ, ÔË ‚ˉËÏ˚ı ÔÓ‚ÂʉÂÌËflı ÔË·Ó‡ ËÎË Ó„‡Ì˘ÂÌÌÓÏ ‚ÓÒÔÓËÁ‚‰ÂÌËË Â„Ó ÙÛÌ͈ËÈ. èÓÎÌÓ ÓÚÍβ˜ÂÌË ÔË·Ó‡ ÓÚ ÒÂÚË „‡‡ÌÚËÛÂÚÒfl ÚÓθÍÓ ÔË ÓÚÒÓ‰ËÌÂÌËË ÒÂÚÂ‚Ó„Ó Í‡·ÂÎfl ÓÚ ÓÁÂÚÍË. ᇢËÚ‡ „·Á ëΉÛÂÚ ËÁ·Â„‡Ú¸ ÔflÏÓ„Ó ËÎË ÌÂÔflÏÓ„Ó ÔÓÔ‡‰‡ÌËfl ËÁÎÛ˜ÂÌËfl ‚ „·Á‡. ÑÎËÚÂθÌÓ ÓÒ‚Â˘ÂÌË ÌÂÔËflÚÌÓ ‰Îfl „·Á Ë ÏÓÊÂÚ Ì‡ÌÂÒÚË ËÏ ‚‰. èÓ˝ÚÓÏÛ ÂÍÓÏẨÛÂÚÒfl ËÒÔÓθÁÓ‚‡Ú¸ Á‡˘ËÚÌ˚È ˝Í‡Ì, ÔÓÒÚ‡‚ÎflÂÏ˚È ‚ÏÂÒÚÂ Ò ‡ÔÔ‡‡ÚÓÏ. ãˈ‡, ÍÓÚÓ˚ ‚ÓÓ·˘Â ˜Û‚ÒÚ‚ËÚÂθÌ˚ Í Ò‚ÂÚÛ, ÔËÌËχ˛Ú ωË͇ÏÂÌÚ˚ ÔÓÚË‚ Ò‚ÂÚÓ˜Û‚ÒÚ‚ËÚÂθÌÓÒÚË ËÎË ÙÓÚÓ˜Û‚ÒÚ‚ËÚÂθÌ˚ Ï‰Ë͇ÏÂÌÚ˚, ÔÂÂÌÂүˠÓÔÂ‡ˆË˛ ̇ „·Á‡, ËÎË ‡·ÓÚ‡˛˘Ë ‚ Ú˜ÂÌË ‰ÎËÚÂθÌÓ„Ó ‚ÂÏÂÌË Ò ˝ÚËÏ ‡ÔÔ‡‡ÚÓÏ ËÎË ‚ ÌÂÔÓÒ‰ÒÚ‚ÂÌÌÓÈ ·ÎËÁÓÒÚË ÓÚ Ì„Ó, Ì ‰ÓÎÊÌ˚ ÔÓ‰‚Â„‡Ú¸Òfl ‰ÂÈÒڂ˲ ˝ÚÓ„Ó ‡ÔÔ‡‡Ú‡, ‰ÓÎÊÌ˚ ÌÓÒËÚ¸ Á‡˘ËÚÌ˚ (Ó‡ÌÊ‚˚Â) Ó˜ÍË, ÍÓÚÓ˚ ‡·ÒÓ·ËÛ˛Ú Ò‚ÂÚ Ò ‰ÎËÌÓÈ ‚ÓÎÌ˚ ‰Ó 515 ÌÏ. ÄÍÍÛÏÛÎflÚÓ ÇÌËχÌËÂ: ËÒÔÓθÁÓ‚‡Ú¸ ÚÓθÍÓ ÓË„Ë̇θÌ˚ ‰ÂÚ‡ÎË – ÓÒÓ·ÂÌÌÓ ˝ÚÓ Í‡Ò‡ÂÚÒfl ‡ÍÍÛÏÛÎflÚÓ‡ Ë Á‡fl‰ÌÓ„Ó ·ÎÓ͇ ÔÓËÁ‚Ó‰ÒÚ‚‡ Ivoclar Vivadent. ÄÍÍÛÏÛÎflÚÓ Ì Á‡ÍÓ‡˜Ë‚‡Ú¸. ç ı‡ÌËÚ¸ Â„Ó ÔË ÚÂÏÔÂ‡ÚÛ ‚˚¯Â 40 °C (ËÎË Í‡ÚÍÓ‚ÂÏÂÌÌÓ ÔË 60°C). ï‡ÌËÚ¸ ÚÓθÍÓ ‚ Á‡flÊÂÌÌÓÏ ÒÓÒÚÓflÌËË. ÑÎËÚÂθÌÓÒÚ¸ ı‡ÌÂÌËfl Ì ‰ÓÎÊ̇ Ô‚˚¯‡Ú¸ 6 ÏÂÒflˆÂ‚. ÇÁ˚‚ÓÓÔ‡ÒÂÌ ÔË ÛÚËÎËÁ‡ˆËË ‚ ÓÚÍ˚ÚÓÏ Ô·ÏÂÌË. é·‡ÚËÚ ‚ÌËχÌËÂ, ˜ÚÓ ÂÒÎË ÎËÚËÈ-ÔÓÎËÏÂÌ˚È ‡ÍÍÛÏÛÎflÚÓ ËÒÔÓθÁÛÂÚÒfl Ì ‚ ÒÓÓÚ‚ÂÚÒÚ‚ËË Ò Ì‡Á̇˜ÂÌËÂÏ Ë ËÌÒÚÛ͈ËÂÈ ËÎË ËÏÂÂÚ ÏÂı‡Ì˘ÂÒÍË ÔÓ‚ÂʉÂÌËfl, ÂÒÚ¸ ‚ÂÓflÚÌÓÒÚ¸ ‚Á˚‚‡, ‚ÓÒÔ·ÏÂÌÂÌËfl ËÎË ÔÓfl‚ÎÂÌËfl ‰˚χ. èÓ‚ÂʉÂÌÌ˚ ÎËÚËÈ-ÔÓÎËÏÂÌ˚ ‡ÍÍÛÏÛÎflÚÓ˚ ËÒÔÓθÁÓ‚‡Ú¸ ÌÂθÁfl. ùÎÂÍÚÓÎËÚ˚, ˝ÎÂÍÚÓÎËÚ˘ÂÒÍË ԇ˚, Ó·‡ÁÛ˛˘ËÂÒfl ÔË ‚Á˚‚Â, ‚ÓÒÔ·ÏÂÌÂÌËË ËÎË Á‡‰˚ÏÎÂÌËË, Ó͇Á˚‚‡˛Ú ÚÓÍÒ˘ÂÒÍÓÂ Ë ‡Á˙‰‡˛˘Â ‰ÂÈÒÚ‚ËÂ. èË ÔÓÔ‡‰‡ÌËË ‚ „·Á‡ Ë Ì‡ ÍÓÊÛ Ò‡ÁÛ Ê ÔÓÏ˚Ú¸ ·Óθ¯ËÏ ÍÓ΢ÂÒÚ‚ÓÏ ‚Ó‰˚. àÁ·Â„‡ÈÚ ‚‰˚ı‡ÌËfl Ô‡Ó‚. èË ÛıÛ‰¯ÂÌËË Ò‡ÏÓ˜Û‚ÒÚ‚Ëfl Ó·‡ÚËÚÂÒ¸ Í ‚‡˜Û. Ç˚‰ÂÎÂÌË ÚÂÔ· Ç ‡·ÓÚ β·Ó„Ó ÏÓ˘ÌÓ„Ó ÔÓÎËÏÂËÁ‡ˆËÓÌÌÓ„Ó ÔË·Ó‡ ËÁÎÛ˜ÂÌË ҂ÂÚ‡ ‚˚ÒÓÍÓÈ ËÌÚÂÌÒË‚ÌÓÒÚË ‚Ò„‰‡ ÒÓÔÓ‚Óʉ‡ÂÚÒfl ‚˚‰ÂÎÂÌËÂÏ ÚÂÔ·. èË ‰ÎËÚÂθÌÓÏ ÓÒ‚Â˘ÂÌËË ÔÛθÔ˚ ËÎË Ïfl„ÍËı Ú͇ÌÂÈ ËÏ ÏÓÊÂÚ ·˚Ú¸ ̇ÌÂÒÂÌ ‚‰ Ò ÌÂÓ·‡ÚËÏ˚ÏË ËÎË Ó·‡ÚËÏ˚ÏË ÔÓÒΉÒÚ‚ËflÏË. èÓ˝ÚÓÏÛ ÒΉÛÂÚ Òӷβ‰‡Ú¸ Ô‰ÔËÒ‡ÌÌÓ ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË. çÂÔÂÏÂÌÌÓ ÒΉÛÂÚ ËÁ·Â„‡Ú¸ ÌÂÔÂ˚‚ÌÓ„Ó ÓÒ‚Â˘ÂÌËfl ‚ Ú˜ÂÌË ·ÓΠ˜ÂÏ 20 ÒÂÍÛ̉ Ó‰ÌÓÈ Ë ÚÓÈ Ê ÔÓ‚ÂıÌÓÒÚË ÁÛ·‡, ‡ Ú‡ÍÊ ÔflÏÓ„Ó ÍÓÌÚ‡ÍÚ‡ Ò ‰ÂÒÌÓÈ, ÒÎËÁËÒÚÓÈ Ó·ÓÎÓ˜ÍÓÈ ËÎË ÍÓÊÂÈ. Ç ÒÎÛ˜‡Â ‚˚ÔÓÎÌÂÌËfl ÂÒÚ‡‚‡ˆËË ÌÂÔflÏ˚Ï ÏÂÚÓ‰ÓÏ ÒΉÛÂÚ ‡·ÓÚ‡Ú¸ Ò ÔÂÂ˚‚‡ÏË, ‚ÂÏÂÌÌÓÈ ËÌÚÂ‚‡Î ‰ÓÎÊÂÌ ÒÓÒÚ‡‚ÎflÚ¸ 20 ÒÂÍÛ̉, ËÎË ÔË Ì‡ÛÊÌÓÏ Óı·ʉÂÌËË Ò ÔÓÏÓ˘¸˛ ÒÚÛË ‚ÓÁ‰Ûı‡. 77 Ç‚Ó‰ ‚ ˝ÍÒÔÎÛ‡Ú‡ˆË˛ èÓ‚Â¸Ú ÍÓÏÔÎÂÍÚÌÓÒÚ¸ ÔÓÒÚ‡‚ÍË (ÒÏ. é·˙ÂÏ ÔÓÒÚ‡‚ÍË) Ë Ì‡Î˘Ë Ú‡ÌÒÔÓÚÌ˚ı ÔÓ‚ÂʉÂÌËÈ. Ç ÒÎÛ˜‡Â, ÂÒÎË Í‡ÍËÂ-ÎË·Ó ˜‡ÒÚË ÔÓ‚ÂʉÂÌ˚ ËÎË ÓÚÒÛÚÒÚ‚Û˛Ú, ÌÂωÎÂÌÌÓ Ò‚flÊËÚÂÒ¸ Ò ÔÓ‰‡‚ˆÓÏ ËÎË ÒÂ‚ËÒÌÓÈ ÒÎÛÊ·ÓÈ. ç‡ÍÓ̘ÌËÍ èÂ‰ ÏÓÌÚ‡ÊÓÏ Ò‚ÂÚÓ‚Ó‰‡ ÒΉÛÂÚ Û‰‡ÎËÚ¸ Á‡˘ËÚÌÛ˛ ÔÎÂÌÍÛ Ò ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘Â„Ó ÓÚ‚ÂÒÚËfl ̇ ̇ÍÓ̘ÌËÍÂ. á‡fl‰Ì˚È ·ÎÓÍ èÂ‰ ‚Íβ˜ÂÌËÂÏ ÒΉÛÂÚ Û‰ÓÒÚÓ‚ÂËÚ¸Òfl, ˜ÚÓ Û͇Á‡ÌÌÓ ̇ Ú‡·Î˘Í ̇ÔflÊÂÌË ÒÓÓÚ‚ÂÚÒÚ‚ÛÂÚ Ì‡ÔflÊÂÌ˲ ‚ ÒÂÚË. 퇷΢͇ ̇ıÓ‰ËÚÒfl ̇ ÌËÊÌÂÈ ˜‡ÒÚË Á‡fl‰ÌÓ„Ó ·ÎÓ͇. ë‚ÂÚÓ‚Ó‰ ‚ÒÚ‡‚ËÚ¸ ‚ ̇ÍÓ̘ÌËÍ Î„ÍËÏ ‚‡˘ÂÌËÂÏ. á‡ÚÂÏ ÛÒÚ‡ÌÓ‚ËÚ¸ ̇ Ò‚ÂÚÓ‚Ó‰ Á‡˘ËÚÌ˚È ˝Í‡Ì. èÓÒÚ‡‚¸Ú Á‡fl‰Ì˚È ·ÎÓÍ Ì‡ ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘Û˛ Ó‚ÌÛ˛ ÔÓ‚ÂıÌÓÒÚ¸ ÒÚÓ·. 쉇ÎËÚ Á‡˘ËÚÌÛ˛ ÔÎÂÌÍÛ Ò ÒÓ‰ËÌËÚÂθÌÓ„Ó ¯ÚÂÍÂ‡ ÒÂÚÂ‚Ó„Ó ·ÎÓ͇. ëÓ‰ËÌËÚÂθÌ˚È ¯ÚÂÍÂ ÒÂÚÂ‚Ó„Ó ·ÎÓ͇ ‚ÒÚ‡‚ËÚ¸ ÔÓ‰ Û„ÎÓÏ ‚ Á‡fl‰Ì˚È ·ÎÓÍ, ‡ Á‡ÚÂÏ Ò Î„ÍËÏ ‰‡‚ÎÂÌË Á‡ÙËÍÒËÓ‚‡Ú¸ (Ú‡Í ˜ÚÓ·˚ ˜Û‚ÒÚ‚Ó‚‡ÎÒfl Ë ·˚Î ÒÎ˚¯ÂÌ ˘ÂΘÓÍ). èÓ‰ÒÓ‰ËÌËÚ¸ ÒÂÚ‚ÓÈ Í‡·Âθ Í ÒÂÚË Ë ÒÂÚ‚ÓÏÛ ·ÎÓÍÛ. é·ÓÁ̇˜ÂÌË Power ON c ΂ÓÈ ÒÚÓÓÌ˚ ÍÓÔÛÒ‡ Á‡„ÓËÚÒfl ÒËÌËÏ ˆ‚ÂÚÓÏ (ÒÏ. ‡Á‰ÂÎ é·ÓÁ̇˜ÂÌËfl ̇ Á‡fl‰ÌÓÏ ·ÎÓÍÂ). 78 ÄÍÍÛÏÛÎflÚÓ èÂ‰ ‚‚Ó‰ÓÏ ‚ ˝ÍÒÔÎÛ‡Ú‡ˆË˛ ‡ÍÍÛÏÛÎflÚÓ ‰ÓÎÊÂÌ ·˚Ú¸ ÔÓÎÌÓÒÚ¸˛ Á‡flÊÂÌ! èÓÎÌÓÒÚ¸˛ Á‡flÊÂÌÌÓ„Ó ‡ÍÍÛÏÛÎflÚÓ‡ ı‚‡Ú‡ÂÚ ÔËÏÂÌÓ Ì‡ 60 ÏËÌÛÚ ÔÓÎËÏÂËÁ‡ˆËË. ìÓ‚Â̸ Á‡fl‰ÍË ‡ÍÍÛÏÛÎflÚÓ‡ ìÓ‚Â̸ Á‡fl‰ÍË ‡ÍÍÛÏÛÎflÚÓ‡ ÓÚÓ·‡Ê‡ÂÚÒfl ̇ ‰ËÒÔΠÔË ‚Íβ˜ÂÌÌÓÏ Ì‡ÍÓ̘ÌËÍ ÒÎÂ‰Û˛˘ËÏ Ó·‡ÁÓÏ: èÓÎÌÓÒÚ¸˛ Á‡flÊÂÌÌ˚È ‡ÍÍÛÏÛÎflÚÓ (‡·Ó˜Â ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË ÒÓÒÚ‡‚ÎflÂÚ ÔËÏÂÌÓ 60 ÏËÌÛÚ). ç‡ÔÓÎÓ‚ËÌÛ Á‡flÊÂÌÌ˚È ‡ÍÍÛÏÛÎflÚÓ ÄÍÍÛÏÛÎflÚÓ ‚ÒÚ‡‚ËÚ¸ ‚ ̇ÍÓ̘ÌËÍ ÔÓ ÔflÏÓÈ, Ú‡Í, ˜ÚÓ·˚ ˜Û‚ÒÚ‚Ó‚‡ÎÒfl Ë ÒÎ˚¯‡ÎÒfl ˘ÂΘÓÍ. êÂÁÂ‚ (Ç ÂÊËÏ êÂÁÂ‚ ÔÓÒΉÌflfl ·‡Î͇ ̇ ‰Ë‡„‡ÏÏ Á‡fl‰ÍË ÒÚ‡ÌÓ‚ËÚÒfl Í‡ÒÌÓÈ. ÄÍÍÛÏÛÎflÚÓ ÒΉÛÂÚ Á‡fl‰ËÚ¸ Í‡Í ÏÓÊÌÓ ÒÍÓÂÂ). ÖÒÎË ‡ÍÍÛÏÛÎflÚÓ Á‡flʇÂÚÒfl, ÔÂʉ ˜ÂÏ ‰ËÒÔÎÂÈ ÔÂÂȉÂÚ ‚ ÂÊËÏ Standby, ̇ ‰ËÒÔΠ‰ÓÎÊ̇ ÔÓfl‚ËÚ¸Òfl Í‡ÚÍÓ‚ÂÏÂÌ̇fl ̇‰ÔËÒ¸ “Charging Battery“. ç ÔËÏÂÌflfl ÛÒËÎËÂ, ‚ÒÚ‡‚ËÚ¸ ̇ÍÓ̘ÌËÍ ‚ ÓÚ‚ÂÒÚË ̇ Á‡fl‰ÌÓÏ ·ÎÓÍÂ. ÖÒÎË ÔËÏÂÌflÂÚÒfl „Ë„ËÂÌ˘ÂÒ͇fl Á‡˘ËÚ‡,  ÒΉÛÂÚ Û‰‡ÎËÚ¸ ÔÂ‰ Á‡fl‰ÍÓÈ ‡ÍÍÛÏÛÎflÚÓ‡. èÓ ‚ÓÁÏÓÊÌÓÒÚË ÔË·Ó ÔÓÒÚÓflÌÌÓ ËÒÔÓθÁÓ‚‡Ú¸ Ò ÔÓÎÌÓÒÚ¸˛ Á‡flÊÂÌÌ˚Ï ‡ÍÍÛÏÛÎflÚÓÓÏ – ˝ÚÓ Ó·ÂÒÔ˜˂‡ÂÚ ‰ÎËÚÂθÌ˚È ÒÓÍ ÒÎÛÊ·˚. èÓ˝ÚÓÏÛ ÂÍÓÏẨÛÂÚÒfl ÔÓÒÎÂ Í‡Ê‰Ó„Ó Ô‡ˆËÂÌÚ‡ ‚ÒÚ‡‚ÎflÚ¸ ̇ÍÓ̘ÌËÍ ‚ Á‡fl‰Ì˚È ·ÎÓÍ. á‡fl‰Í‡ ÔÓÎÌÓÒÚ¸˛ ‡Áfl‰Ë‚¯Â„ÓÒfl ‡ÍÍÛÏÛÎflÚÓ‡ ‰ÎËÚÒfl 2 ˜‡Ò‡. èË ÔÓÎÌÓÒÚ¸˛ ‡ÁflÊÂÌÌÓÏ ‡ÍÍÛÏÛÎflÚÓ ̇ÍÓ̘ÌËÍ ‡‚ÚÓχÚ˘ÂÒÍË ÔÂÂÍβ˜‡ÂÚÒfl ‚ ÒÓÒÚÓflÌË Stand-by. àÁÎÛ˜ÂÌË ·Óθ¯Â Ì ‡ÍÚË‚ËÛÂÚÒfl, ÔÓ„‡ÏÏÛ Ë ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË Á‡‰‡Ú¸ Ì‚ÓÁÏÓÊÌÓ. é‰Ì‡ÍÓ, ‚ Ú‡ÍËı ÒÎÛ˜‡flı ̇ÍÓ̘ÌËÍÓÏ ÏÓÊÌÓ ‡·ÓÚ‡Ú¸, ËÒÔÓθÁÛfl ÙÛÌÍˆË˛ ‡·ÓÚ˚ ˜ÂÂÁ ͇·Âθ Click & Cure. ÄÍÍÛÏÛÎflÚÓ fl‚ÎflÂÚÒfl ·˚ÒÚÓ ËÁ̇¯Ë‚‡˛˘ÂÈÒfl ˜‡ÒÚ¸˛, Ó·˚˜ÌÓ ˜ÂÂÁ 2 Ò ÔÓÎÓ‚ËÌÓÈ „Ó‰‡ ÓÌ ÔÓ‰ÎÂÊËÚ Á‡ÏÂÌÂ. «ÇÓÁ‡ÒÚ» ‡ÍÍÛÏÛÎflÚÓ‡ ÏÓÊÌÓ ÛÁ̇ڸ ÔÓ Ì‡ÍÎÂÈÍ ̇ ÌÂÏ. 79 àÒÔÓθÁÓ‚‡ÌË ÔË·Ó‡ ꇷÓÚ‡ ˜ÂÂÁ ͇·Âθ Click & Cure ã‡ÏÔÓÈ Bluephase N ‚ β·Ó ‚ÂÏfl ÏÓÊÌÓ ‡·ÓÚ‡Ú¸ ˜ÂÂÁ ͇·Âθ, ‡ ÓÒÓ·ÂÌÌÓ ÔË ÔÓÎÌÓÒÚ¸˛ ‡ÁflÊÂÌÌÓÏ ‡ÍÍÛÏÛÎflÚÓÂ. ÑÎfl ˝ÚÓ„Ó ÌÂÓ·ıÓ‰ËÏÓ ‚˚ÌÛÚ¸ ‡ÍÍÛÏÛÎflÚÓ ËÁ ̇ÍÓ̘ÌË͇. èÓÒΠ˝ÚÓ„Ó ÓÚÒÓ‰ËÌËÚ¸ ¯ÚÂÍÂ ÒÂÚÂ‚Ó„Ó ·ÎÓ͇ ÔËÚ‡ÌËfl Ò ÌËÊÌÂÈ ÒÚÓÓÌ˚ Á‡fl‰ÌÓ„Ó ·ÎÓ͇. èË ˝ÚÓÏ Ì ÒΉÛÂÚ ÚflÌÛÚ¸ Á‡ ÒÂÚ‚ÓÈ Í‡·Âθ. èÂ‰ ͇ʉ˚Ï ËÒÔÓθÁÓ‚‡ÌËÂÏ Á‡„flÁÌÂÌÌ˚ ÔÓ‚ÂıÌÓÒÚË ÔË·Ó‡, ‡ Ú‡ÍÊ ҂ÂÚÓ‚Ó‰ Ë Á‡˘ËÚÌ˚È ÍÓÁ˚ÂÍ ÒΉÛÂÚ ‰ÂÁËÌÙˈËÓ‚‡Ú¸ ËÎË ‡‚ÚÓÍ·‚ËÓ‚‡Ú¸. äÓÏ ÚÓ„Ó, ÒΉÛÂÚ Û‰ÓÒÚÓ‚ÂËÚ¸Òfl, ˜ÚÓ ÏÓ˘ÌÓÒÚ¸ Ò‚ÂÚ‡ ÔË·Ó‡ ‰ÓÒÚ‡ÚӘ̇ ‰Îfl ÔÓÎËÏÂËÁ‡ˆËË Ï‡ÚÂˇ·. ÑÎfl ˝ÚÓ„Ó ÒΉÛÂÚ „ÛÎflÌÓ ÔÓ‚ÂflÚ¸ Ò‚ÂÚÓ‚Ó‰ ̇ Á‡„flÁÌÂÌËfl Ë ÔÓ‚ÂʉÂÌËfl, ‡ Ú‡ÍÊ ÍÓÌÚÓÎËÓ‚‡Ú¸ ÏÓ˘ÌÓÒÚ¸ ËÁÎÛ˜ÂÌËfl (̇ÔËÏÂ, Ò ÔÓÏÓ˘¸˛ ‡‰ËÓÏÂÚ‡ “Bluephase Meter“ ÙËÏ˚ Ivoclar Vivadent). Ç˚·Ó ÔÓ„‡ÏÏ˚ Ë ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË èÓ„‡ÏÏÛ Ë ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË ÏÓÊÌÓ Ì‡ÒÚ‡Ë‚‡Ú¸ Ë̉˂ˉۇθÌÓ. ÑÎfl ‡Á΢Ì˚ı ÔÓ͇Á‡ÌËÈ Î‡ÏÔ‡ Bluephase N ËÏÂÂÚ 3 ÔÓ„‡ÏÏ˚ ÔÓÎËÏÂËÁ‡ˆËË. ë ÔÓÏÓ˘¸˛ Í·‚Ë¯Ë ‚˚·Ó‡ ÔÓ„‡ÏÏ˚ ÏÓÊÌÓ Á‡‰‡Ú¸ Ê·ÂÏÛ˛ ÔÓ„‡ÏÏÛ ÔÓÎËÏÂËÁ‡ˆËË. àÁÓ·‡ÊÂÌË ̇ ‰ËÒÔΠ·Û‰ÂÚ ËÁÏÂÌflÚ¸Òfl ‚ ÒÓÓÚ‚ÂÚÒÚ‚ËË Ò ‚˚·ÓÓÏ (ÒÏ. é·ÓÁ̇˜ÂÌËfl ̇ ̇ÍÓ̘ÌËÍÂ). ë Á‡‚Ó‰‡ ÔË·Ó ÔÓÒÚ‡‚ÎflÂÚÒfl ÒÓ ÒÎÂ‰Û˛˘ËÏË Ì‡ÒÚÓÈ͇ÏË: – HIP (High Power – ‚˚ÒÓ͇fl ÏÓ˘ÌÓÒÚ¸): 10 ÒÂÍÛ̉ – LOW (Low Power – ÌËÁ͇fl ÏÓ˘ÌÓÒÚ¸): 10 ÒÂÍÛ̉ – SOFí (Soft start – Ïfl„ÍËÈ ÒÚ‡Ú): 15 ÒÂÍÛ̉ Ü·ÂÏÓ ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË Á‡‰‡ÂÚÒfl Í·‚˯ÂÈ ‚˚·Ó‡ ‚ÂÏÂÌË. åÓÊÌÓ ‚˚·‡Ú¸ 5, 10, 15, 20 Ë 30 ÒÂÍÛ̉. òÚÂÍÂ ‚ÒÚ‡‚ËÚ¸ ‚ ̇ÍÓ̘ÌËÍ Ú‡ÍËÏ Ó·‡ÁÓÏ, ˜ÚÓ·˚ ˜Û‚ÒÚ‚Ó‚‡ÎÒfl Ë ·˚Î ÒÎ˚¯ÂÌ ˘ÂΘÓÍ. ÇÓ ‚ÂÏfl ‡·ÓÚ˚ ˜ÂÂÁ ͇·Âθ Ì‚ÓÁÏÓÊÌÓ Á‡flʇڸ ‡ÍÍÛÏÛÎflÚÓ Ò ÔÓÏÓ˘¸˛ Á‡fl‰ÌÓ„Ó ·ÎÓ͇ ÔÓ Ô˘ËÌ ÓÚÒÛÚÒÚ‚Ëfl ˝ÌÂ„ÓÒ̇·ÊÂÌËfl. èÓÎÌÓ ÓÚÍβ˜ÂÌË ÓÚ ÒÂÚË „‡‡ÌÚËÛÂÚÒfl ÚÓθÍÓ ÔË ‚˚ÌÛÚÓÏ ËÁ ÓÁÂÚÍË ÒÂÚ‚ÓÏ Í‡·ÂÎÂ. 80 èË ‚˚·Ó ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË ÒΉÛÂÚ Û˜ËÚ˚‚‡Ú¸ Ú·ӂ‡ÌËfl ËÌÒÚÛ͈ËË Í ËÒÔÓθÁÛÂÏÓÏÛ Ï‡ÚÂˇÎÛ. èË ‡·ÓÚÂ Ò ÍÓÏÔÓÁËÚ‡ÏË ÂÍÓÏẨ‡ˆËË Ó ÔÓÎËÏÂËÁ‡ˆËË ‡ÒÔÓÒÚ‡Ìfl˛ÚÒfl ̇ ‚Ò ˆ‚ÂÚ‡ Ë ÒÎÓË Ï‡ÚÂˇ· ÚÓ΢ËÌÓÈ Ì ·ÓΠ2 ÏÏ – ÂÒÎË ËÌÓ Ì Û͇Á‡ÌÓ ‚ ËÌÒÚÛ͈ËË. ùÚË ÂÍÓÏẨ‡ˆËË ‰ÂÈÒÚ‚ËÚÂθÌ˚ ‚ ÚÓÏ ÒÎÛ˜‡Â, ÍÓ„‰‡ ÔÓÎËÏÂËÁ‡ˆËfl ÓÒÛ˘ÂÒÚ‚ÎflÂÚÒfl Ú‡ÍËÏ Ó·‡ÁÓÏ, ˜ÚÓ ‚˚ıÓ‰ÌÓ ÓÍÓ¯ÍÓ Ò‚ÂÚÓ‚Ó‰‡ ‡ÒÔÓ·„‡ÂÚÒfl ÔflÏÓ ÔÂ‰ ÓÚ‚Âʉ‡ÂÏ˚Ï Ï‡ÚÂˇÎÓÏ. ë Û‚Â΢ÂÌËÂÏ ‡ÒÒÚÓflÌËfl ÏÂÊ‰Û ÌËÏË ‰ÓÎÊÌÓ ÒÓÓÚ‚ÂÚÒÚ‚ÂÌÌÓ Û‚Â΢˂‡Ú¸Òfl Ë ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË. ÖÒÎË ‡ÒÒÚÓflÌË ÏÂÊ‰Û Ò‚ÂÚÓ‚Ó‰ÓÏ Ë Ï‡ÚÂˇÎÓÏ ÒÓÒÚ‡‚ÎflÂÚ 9 ÏÏ, ˝ÙÙÂÍÚ˂̇fl ÏÓ˘ÌÓÒÚ¸ ËÁÎÛ˜ÂÌËfl ÒÌËʇÂÚÒfl ÔËÏÂÌÓ Ì‡ 50%, Ú‡Í ˜ÚÓ ÂÍÓÏẨÛÂÏÓ ‚ÂÏfl ÔÓÎËÏÂËÁ‡ˆËË ‚ ˝ÚÓÏ ÒÎÛ˜‡Â ÒΉÛÂÚ Û‰‚ÓËÚ¸. èÓ„‡Ïχ ÔÓÎËÏÂËÁ‡ˆËË èÓ„‡Ïχ HIGH POWER èÓÒÚÓflÌÌÓ ‚˚ÒÓ͇fl ÏÓ˘ÌÓÒÚ¸ Ò‚ÂÚ‡ ‰Îfl ÔÓÎËÏÂËÁ‡ˆËË ÔÎÓÏ·ËÓ‚Ó˜Ì˚ı Ë ÙËÍÒËÛ˛˘Ëı χÚÂˇÎÓ‚ ÔË ‚˚ÔÓÎÌÂÌËË ÂÒÚ‡‚‡ˆËÈ ÔflÏ˚Ï Ë ÌÂÔflÏ˚Ï ÏÂÚÓ‰ÓÏ. 1.200 mW/cm2 0 t [s] èÓ„‡Ïχ LOW POWER èÓÌËÊÂÌ̇fl ÏÓ˘ÌÓÒÚ¸ Ò‚ÂÚ‡ Ò ÛÏÂ̸¯ÂÌÌ˚Ï ‚˚‰ÂÎÂÌËÂÏ ÚÂÔ· ‰Îfl ÔÓÎËÏÂËÁ‡ˆËË ‚ ӷ·ÒÚflı, ·ÎËÁÍËı Í ÔÛθÔÂ, ‡‰„ÂÁË‚Ó‚, ·ÈÌÂÓ‚ Ë ÔÎÓÏ·ËÓ‚Ó˜Ì˚ı χÚÂˇÎÓ‚ ÔË ÂÒÚ‡‚‡ˆËflı ÔÓÎÓÒÚÂÈ Í·ÒÒ‡ V. 650 mW/cm2 0 èÎÓÏ·ËÓ‚Ó˜Ì˚ χÚÂˇÎ˚ äÓÏÔÓÁËÚ˚ • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow ‚Ò Í·ÒÒ˘ÂÒÍË ÍÓÏÔÓÁËÚ˚ • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White äÓÏÔÓÏÂ 3) Compoglass F / Compoglass Flow çÂÔflÏ˚ ÂÒÚ‡‚‡ˆËË/ îËÍÒËÛ˛˘Ë χÚÂˇÎ˚ Variolink II Base 4) / Variolink Veneer / Variolink N Base 4) / Variolink N Clear Veneer / Variolink Ultra 5) / Dual Cement 5) / Variolink II 5) èӘˠχÚÂˇÎ˚ Helioseal / Helioseal F / Helioseal Clear Monopaque MultiCore Flow / Multicore HB Heliosit Orthodontic Telio Add-On Flow Telio Stains IPS Empress Direct Color IPS Empress Direct Opaque ĉ„ÂÁË‚˚ AdheSE / AdheSE One F ExciTE / ExciTE DSC Tetric N-Bond / Tetric N-Bond Self-Etch Heliobond (Syntac) ÇÂÏÂÌÌ˚ χÚÂˇÎ˚ Telio CS Inlay/Onlay Systemp.inlay/onlay Fermit / Fermit N Telio CS Link / Systemp.link èӘˠχÚÂˇÎ˚ Heliosit Orthodontic Vivaglass Liner ÇÂÏfl ÔÓÎËÏÂËÁ‡ˆËË 10 ÒÂÍ. 15 ÒÂÍ. 10 ÒÂÍ. 20 ÒÂÍ. ç‡ ÏÏ ÍÂ‡ÏËÍË: 10 ÒÂÍ. ̇ Í‡Ê‰Û˛ ÔÓ‚ÂıÌÓÒÚ¸ 10 ÒÂÍ. 20 ÒÂÍ. 20 ÒÂÍ. 10 ÒÂÍ. 15 ÒÂÍ. 10 ÒÂÍ. 10 ÒÂÍ. 20 ÒÂÍ. ÇÂÏfl ÔÓÎËÏÂËÁ‡ˆËË 10 ÒÂÍ. 10 ÒÂÍ. 10 ÒÂÍ. 10 ÒÂÍ. 20 ÒÂÍ. ̇ Í‡Ê‰Û˛ ÔÓ‚ÂıÌÓÒÚ¸ 20 ÒÂÍ. 20 ÒÂÍ. t [s] 1.200 mW/cm2 èÓ„‡Ïχ SOFT START ëÚÛÔÂ̘‡ÚÓ ÔÓ‚˚¯ÂÌË ÏÓ˘ÌÓÒÚË Ò‚ÂÚ‡, ÒÌËÊÂÌË ÒÚÂÒÒ‡ χÚÂˇ· ÔË ÛÒ‡‰ÍÂ, ÛÏÂ̸¯ÂÌÌÓ ‚˚‰ÂÎÂÌË ÚÂÔ· ‰Îfl ÔÓÎËÏÂËÁ‡ˆËË ÔÎÓÏ·ËÓ‚Ó˜Ì˚ı χÚÂˇÎÓ‚. 650 mW/cm 2 0 1,200 mW/cm2 åÓ˘ÌÓÒÚ¸ ËÁÎÛ˜ÂÌËfl 15 ÒÂÍ. ‚ÂÏfl ÓÚ‚ÂʉÂÌËfl ‰Îfl ÍÓÏÔÓÁËÚÓ‚ ‚ÂÏfl ÓÚ‚ÂʉÂÌËfl ‰Îfl Tetric EvoCeram/IPS Empress Direct 10 ÒÂÍ. 5 èÎÓÏ·ËÓ‚Ó˜Ì˚ χÚÂˇÎ˚ äÓÏÔÓÁËÚ˚ • 2 mm 1) IPS Empress Direct / IPS Empress Direct Flow / Tetric EvoCeram / Tetric EvoFlow / Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow Heliomolar / Heliomolar HB / Heliomolar Flow ‚Ò Í·ÒÒ˘ÂÒÍË ÍÓÏÔÓÁËÚ˚ • 4 mm 2) Tetric EvoCeram Bulk Fill / Tetric Basic White äÓÏÔÓÏÂ 3) Compoglass F / Compoglass Flow ÇÂÏfl ÔÓÎËÏÂËÁ‡ˆËË 15 ÒÂÍ. 20 ÒÂÍ. 15 ÒÂÍ. 20 ÒÂÍ. t [s] 1) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl ÚÓ΢ËÌ˚ ÒÎÓfl Ì ·ÓΠ2 ÏÏ, ‡ Ú‡ÍÊ ÂÒÎË ‚ ËÌÒÚÛ͈ËË Í Ï‡ÚÂˇÎÛ ÌÂÚ ‰Û„Ëı Û͇Á‡ÌËÈ (‚ÓÁÏÓÊÌÓ, ̇ÔËÏÂ, Ò ˆ‚ÂÚ‡ÏË ‰ÂÌÚË̇) 2) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl ÚÓ΢ËÌ˚ ÒÎÓfl Ì ·ÓΠ4 ÏÏ, ‡ Ú‡ÍÊ ÂÒÎË ‚ ËÌÒÚÛ͈ËË Í Ï‡ÚÂˇÎÛ ÌÂÚ ‰Û„Ëı Û͇Á‡ÌËÈ (‚ÓÁÏÓÊÌÓ, ̇ÔËÏÂ, Ò ˆ‚ÂÚ‡ÏË ‰ÂÌÚË̇) 3) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl ÚÓ΢ËÌ˚ ÒÎÓfl Ì ·ÓΠ3 ÏÏ 4) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl Ò‚ÂÚÓ‚ÓÈ ÔÓÎËÏÂËÁ‡ˆËË (ËÒÔÓθÁÓ‚‡ÌË ÚÓθÍÓ Ô‡ÒÚ˚ ·‡Á˚) 5) ‰ÂÈÒÚ‚ËÚÂθÌÓ ‰Îfl ‰‚ÓÈÌÓÈ ÔÓÎËÏÂËÁ‡ˆËË 81 îÛÌ͈Ëfl ÒÓı‡ÌÂÌËfl ‰‡ÌÌ˚ı Cure Memory èÓÒΉÌË ËÒÔÓθÁÓ‚‡ÌÌ˚ ̇ÒÚÓÈÍË ÔÓ„‡ÏÏ˚ ÔÓÎËÏÂËÁ‡ˆËË Ë ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË ‡‚ÚÓχÚ˘ÂÒÍË ÒÓı‡Ìfl˛ÚÒfl. ëÚ‡Ú äÌÓÔÍÓÈ ëÚ‡Ú ‚Íβ˜‡ÂÚÒfl Ò‚ÂÚÓ‚Ó ËÁÎÛ˜ÂÌËÂ. èÓ ËÒÚ˜ÂÌËË Á‡‰‡ÌÌÓ„Ó ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË ÔÓ„‡Ïχ Á‡‚Â¯‡ÂÚÒfl ‡‚ÚÓχÚ˘ÂÒÍË. èË ÌÂÓ·ıÓ‰ËÏÓÒÚË Ò‚ÂÚ ÏÓÊÌÓ ÔÂʉ‚ÂÏÂÌÌÓ ÓÚÍβ˜ËÚ¸ ÔÓ‚ÚÓÌ˚Ï Ì‡Ê‡ÚËÂÏ Ì‡ Í·‚Ë¯Û ëÚ‡Ú. é‰ÌÓ‚ÂÏÂÌÌÓ Ò ‚Íβ˜ÂÌËÂÏ Ò‚ÂÚ‡ ̇˜Ë̇ÂÚ ‡·ÓÚ‡Ú¸ ‚ÂÌÚËÎflÚÓ. èÓ ÓÍÓ̘‡ÌËË ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË ÓÌ ÔÓ‰ÓÎʇÂÚ ‡·ÓÚ‡Ú¸, Óı·ʉ‡fl ÔË·Ó. èÓ͇ ‚ÂÌÚËÎflÚÓ ‡·ÓÚ‡ÂÚ, ‡ÍÍÛÏÛÎflÚÓ ‚˚ÌËχڸ ÌÂθÁfl. á‚ÛÍÓ‚˚ Ò˄̇Î˚ Ç ÒÎÂ‰Û˛˘Ëı ÒÎÛ˜‡flı Á‚Û˜‡Ú Á‚ÛÍÓ‚˚ Ò˄̇Î˚: – ëÚ‡Ú (ëÚÓÔ) – ä‡Ê‰˚ 10 ÒÂÍÛ̉ – ëÏÂ̇ ÔÓ„‡ÏÏ˚ ÔÓÎËÏÂËÁ‡ˆËË – ëÏÂ̇ ‚ÂÏÂÌË ÔÓÎËÏÂËÁ‡ˆËË – ëÓ‰ËÌËÚ¸ ‡ÍÍÛÏÛÎflÚÓ Ò Á‡fl‰Ì˚Ï ·ÎÓÍÓÏ – ìÒÚ‡ÌÓ‚ËÚ¸ ‡ÍÍÛÏÛÎflÚÓ – ëÓÓ·˘ÂÌË ӷ ӯ˷Í èË Ê·ÌËË „ÓÏÍÓÒÚ¸ Á‚ÛÍÓ‚˚ı Ò˄̇ÎÓ‚ ÏÓÊÌÓ „ÛÎËÓ‚‡Ú¸. ÑÎfl ˝ÚÓ„Ó ÔË ‚˚Íβ˜ÂÌÌÓÏ Ò‚ÂÚÓ‚ÓÏ ËÁÎÛ˜ÂÌËË ÒΉÛÂÚ Ì‡Ê‡Ú¸ ̇ ÒËÌ˛˛ Í·‚Ë¯Û „ÓÏÍÓÒÚË (ÒÏ. èÂ˜Â̸ ˜‡ÒÚÂÈ) Ë Î‚ÓÈ Í·‚˯ÂÈ ‚˚·Ó‡ ‚ÂÏÂÌË ÛÏÂ̸¯ËÚ¸ Á‚ÛÍ ËÎË ÒÓ‚ÒÂÏ ÓÚÍβ˜ËÚ¸ Á‚ÛÍÓ‚ÓÈ Ò˄̇Î. ÖÒÎË Ç˚ ıÓÚËÚ ‚Íβ˜ËÚ¸ Á‚ÛÍÓ‚ÓÈ Ò˄̇ΠËÎË Û‚Â΢ËÚ¸ „ÓÏÍÓÒÚ¸, ÒΉÛÂÚ ÒÌÓ‚‡ ̇ʇڸ ̇ ÒËÌ˛˛ Í·‚Ë¯Û „ÓÏÍÓÒÚË Ë Ô‡‚Û˛ Í·‚Ë¯Û ‚˚·Ó‡ ‚ÂÏÂÌË. 82 åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl ÒÓı‡ÌflÂÚÒfl ÔÓÒÚÓflÌÌÓÈ ÔË ‡·ÓÚ ÔË·Ó‡. èË ËÒÔÓθÁÓ‚‡ÌËË ÔÓÒÚ‡‚ÎflÂÏÓ„Ó Ò ÔË·ÓÓÏ Ò‚ÂÚÓ‚Ó‰‡ 10 ÏÏ ÏÓ˘ÌÓÒÚ¸ ÓÚ͇ÎË·Ó‚‡Ì‡ ̇ 1.200 ÏÇÚ/ÒÏ2 ± 10% ÖÒÎË ËÒÔÓθÁÛÂÚÒfl ‰Û„ÓÈ Ò‚ÂÚÓ‚Ó‰, ‡ Ì ÚÓÚ, ˜ÚÓ ÔÓÒÚ‡‚ÎflÂÚÒfl Ò ÔË·ÓÓÏ, ÓÌ ÏÓÊÂÚ Ó͇Á‡Ú¸ ÒÂ¸ÂÁÌÓ ‚ÎËflÌË ̇ ÏÓ˘ÌÓÒÚ¸ ËÁÎÛ˜ÂÌËfl ̇ ‚˚ıÓ‰Â. èË ËÒÔÓθÁÓ‚‡ÌËË Ò‚ÂÚÓ‚Ó‰‡ Ò Ô‡‡ÎÎÂθÌ˚ÏË ÒÚÂÌ͇ÏË (10 ÏÏ) ‰Ë‡ÏÂÚ˚ ‚ıÓ‰fl˘Â„Ó Ë ‚˚ıÓ‰fl˘Â„Ó Ò‚ÂÚÓ‚Ó„Ó ÓÍӯ͇ Ó‰Ë̇ÍÓ‚˚. èË ËÒÔÓθÁÓ‚‡ÌËË ÙÓÍÛÒËÛ˛˘Ëı Ò‚ÂÚÓ‚Ó‰Ó‚ (10>8 ÏÏ, Pin-Point 6>2 ÏÏ) ‰Ë‡ÏÂÚ ̇ ‚ıӉ ·Óθ¯Â, ˜ÂÏ ‰Ë‡ÏÂÚ ̇ ‚˚ıÓ‰Â. èË ˝ÚÓÏ Ì‡ ‚˚ıӉ ҂ÂÚ ÙÓÍÛÒËÛÂÚÒfl ̇ Ì·Óθ¯ÓÈ ÔÓ‚ÂıÌÓÒÚË, ˜ÚÓ ÔÓ‚˚¯‡ÂÚ ÏÓ˘ÌÓÒÚ¸ ËÁÎÛ˜ÂÌËfl. ë‚ÂÚÓ‚Ó‰˚ Pin-Point ıÓÓ¯Ó ÔÓ‰ıÓ‰flÚ ‰Îfl ÚӘ˜ÌÓÈ ÔÓÎËÏÂËÁ‡ˆËË, ̇ÔËÏÂ, ÙËÍÒ‡ˆËË ‚ËÌËÓ‚ ÔÂ‰ Û‰‡ÎÂÌËÂÏ ËÁÎ˯ÍÓ‚. ÑÎfl Ôӂ‰ÂÌËfl ÔÓÎÌÓ„Ó ÓÚ‚ÂʉÂÌËfl Ò‚ÂÚÓ‚Ó‰ ÒΉÛÂÚ Á‡ÏÂÌËÚ¸. àÁÏÂÂÌË ÏÓ˘ÌÓÒÚË Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl ·ÏÔ˚ Bluephase N Ë ÔÓÒÚ‡‚ÎflÂÏÓ„Ó Ò ÌÂÈ Ò‚ÂÚÓ‚Ó‰‡ ̇ 10 ÏÏ ÏÓÊÌÓ ÔÓ‚ÂËÚ¸, ̇ÔËÏÂ, Ò ÔÓÏÓ˘¸˛ Bluephase Meter. ÖÒÎË ËÁÏÂÂÌÌÓ Á̇˜ÂÌË Ì ÒÓÓÚ‚ÂÚÒÚ‚ÛÂÚ ÓÊˉ‡ÂÏÓÏÛ: – ÔÓ‚ÂËÚ¸ ‚˚·‡ÌÌÛ˛ ÔÓ„‡ÏÏÛ ÔÓÎËÏÂËÁ‡ˆËË – Ó˜ËÒÚËÚ¸ ÓÚ Á‡„flÁÌÂÌËÈ Ò‚ÂÚÓ‚ÓÈ ÒÂÌÒÓ ‡‰ËÓÏÂÚ‡ – ÒÌflÚ¸ Ò‚ÂÚÓ‚Ó‰, ‡ ‚˚ıÓ‰ÌÓ ÓÚ‚ÂÒÚË ҂ÂÚ‡ ̇ ̇ÍÓ̘ÌËÍ ÔÓ˜ËÒÚËÚ¸ ‚‡ÚÌÓÈ Ô‡ÎÓ˜ÍÓÈ, ÒÏÓ˜ÂÌÌÓÈ ÒÔËÚÓÏ. – ‚ÓÁÏÓÊÌÓ, Á‡„flÁÌËÎÒfl Ò‚ÂÚÓ‚Ó‰ Ë Â„Ó ÒΉÛÂÚ ÔÓ˜ËÒÚËÚ¸ (ÒÏ. ‡Á‰ÂÎ íÂıÓ·ÒÎÛÊË‚‡ÌËÂ Ë Ó˜ËÒÚ͇) – ÂÒÎË Ò‚ÂÚÓ‚Ó‰ ÔÓ‚ÂʉÂÌ, ÌÂÓ·ıÓ‰ËÏÓ Á‡ÏÂÌËÚ¸ Â„Ó Ì‡ ÌÓ‚˚È ÖÒÎË ˝ÚË ÏÂ˚ Ì ÔËÌÂÒÎË Ê·ÂÏÓ„Ó ÂÁÛθڇڇ, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û. íÂıÓ·ÒÎÛÊË‚‡ÌËÂ Ë Ó˜ËÒÚ͇ ÑÎfl ÎÛ˜¯Â„Ó Òӷβ‰ÂÌËfl Ú·ӂ‡ÌËÈ „Ë„ËÂÌ˚ ÂÍÓÏẨÛÂÚÒfl ‰Îfl Í‡Ê‰Ó„Ó Ô‡ˆËÂÌÚ‡ ËÒÔÓθÁÓ‚‡Ú¸ Ó‰ÌÓ‡ÁÓ‚˚ ˜ÂıÎ˚. ᇄflÁÌÂÌÌ˚ ÔÓ‚ÂıÌÓÒÚË ÔË·Ó‡, ‡ Ú‡ÍÊ Á‡˘ËÚÌ˚È ÍÓÁ˚ÂÍ ÒΉÛÂÚÔÂ‰ ͇ʉ˚Ï ËÒÔÓθÁÓ‚‡ÌËÂÏ ‰ÂÁËÌÙˈËÓ‚‡Ú¸ (̇ÔËÏÂ, FD333, FD366/Dürr Dental, Incidin Liquid/Ecolab), ‚ ÒÎÛ˜‡Â ÂÒÎË Ì ·˚ÎË ËÒÔÓθÁÓ‚‡Ì˚ Á‡˘ËÚÌ˚ ˜ÂıÎ˚ Ò‚ÂÚÓ‚Ó‰ ‰ÓÎÊÂÌ ·˚Ú¸ ÒÚÂÂÎËÁÓ‚‡Ì ÔÂ‰ ͇ʉ˚Ï ÔËÏÂÌÂÌËÂÏ. èË ‡·ÓÚ‡ı ÔÓ Ó˜ËÒÚÍ ÔË·Ó‡ ̇ ̇ÍÓ̘ÌËÍ, Á‡fl‰Ì˚È ·ÎÓÍ, ‡ ‚ ÓÒÓ·ÂÌÌÓÒÚË Ì‡ ÒÂÚ‚ÓÈ ·ÎÓÍ Ì ‰ÓÎÊÌ˚ ÔÓÔ‡‰‡Ú¸ ÌË͇ÍË ÊˉÍÓÒÚË ËÎË ‰Û„Ë χÚÂˇÎ˚ (ÓÔ‡ÒÌÓÒÚ¸ ÔÓ‡ÊÂÌËfl ˝ÎÂÍÚ˘ÂÒÍËÏ ÚÓÍÓÏ). èË Ó˜ËÒÚÍ Á‡fl‰Ì˚È ·ÎÓÍ ÌÂÓ·ıÓ‰ËÏÓ ÓÚÍβ˜ËÚ¸ ÓÚ ˝ÎÂÍÚÓÒÂÚË. ÏÓ„ÛÚ ÛÒËÎËÚ¸ ‰ÂÈÒÚ‚ËÂ). êÂÍÓÏẨÛÂÚÒfl ÌÂÈÚ‡Î¸ÌÓ ÙÂÏÂÌÚ‡ˆËÓÌÌÓ ÏÓ˛˘Â Ò‰ÒÚ‚Ó. é·‡˘‡ÈÚ ‚ÌËχÌËÂ, ˜ÚÓ Ò‰ÒÚ‚Ó, ËÒÔÓθÁÛÂÏÓ ‰Îfl Ó˜ËÒÚÍË Ë ‰ÂÁËÌÙÂÍ-ˆËË Ì ÒÓ‰ÂÊËÚ – Ó„‡Ì˘ÂÒÍËÂ, ÏËÌÂ‡Î¸Ì˚Â Ë ÓÍËÒÎfl˛˘Ë ÍËÒÎÓÚ˚ (ÏËÌËχθÌÓ ‰ÓÔÛÒÚËÏÓ Á̇˜Â-ÌË pH 5,5), – ˘ÂÎÓ˜Ë (χÍÒËχθÌÓ ‰ÓÔÛÒÚËÏÓ Á̇˜ÂÌË pH 8,5), – ÓÍËÒÎfl˛˘Ë Ò‰ÒÚ‚‡ (̇ÔËÏÂ, ÔÂÓÍÒˉ ‚Ó‰ÓÓ‰‡) èÓÒΠ˝ÚÓ„Ó ‚˚̸Ú ҂ÂÚÓ‚Ó‰ ËÁ ‡ÒÚ‚Ó‡ Ë Ú˘‡ÚÂθÌÓ ÔÓÏÓÈÚ ÔÓ‰ ÔÓÚÓ˜ÌÓÈ ‚Ó‰ÓÈ (Ì ÏÂÌ 10 ÒÂÍÛ̉). ùÙÙÂÍÚË‚ÌÓÈ ‡Î¸ÚÂ̇ÚË‚ÓÈ fl‚ÎflÂÚÒfl Ó˜ËÒÚ͇ ‚ ÚÂÏÓ‰ÂÁËÌ-ÙÂÍÚÓÂ. ëÚÂËÎËÁ‡ˆËfl äÓÔÛÒ ç‡ÍÓ̘ÌËÍ Ë ‰ÂʇÚÂθ ̇ÍÓ̘ÌË͇ ÔÓÚË‡Ú¸ Ó·˚˜Ì˚Ï ‰ÂÁËÌÙÂ͈ËÓÌÌ˚Ï Ò‰ÒÚ‚ÓÏ, Ì ÒÓ‰Âʇ˘ËÏ ‡Î¸‰Â„ˉӂ. ç ËÒÔÓθÁÓ‚‡Ú¸ ÌË͇ÍËı ‡„ÂÒÒË‚Ì˚ı ‰ÂÁËÌÙÂ͈ËÓÌÌ˚ı Ò‰ÒÚ‚ (̇ÔËÏÂ, ‡ÒÚ‚Ó˚ ̇ ÓÒÌÓ‚Â ‡ÔÂθÒËÌÓ‚Ó„Ó Ï‡Ò· ËÎË ‡ÒÚ‚Ó˚ Ò ÒÓ‰ÂʇÌËÂÏ ˝Ú‡ÌÓ· Ò‚˚¯Â 40%), ‡ÒÚ‚ÓËÚÂÎÂÈ (̇ÔËÏÂ, ‡ˆÂÚÓ̇) ËÎË ÓÒÚ˚ı Ô‰ÏÂÚÓ‚, ÍÓÚÓ˚ ÏÓ„ÛÚ ÔÓ‚‰ËÚ¸ ËÎË ÔÓˆ‡‡Ô‡Ú¸ Ô·ÒÚχÒÒÛ. ᇄflÁÌÂÌÌ˚ Ô·ÒÚχÒÒÓ‚˚ ˜‡ÒÚË Ó˜Ë˘‡Ú¸ Ï˚θÌ˚Ï ‡ÒÚ‚ÓÓÏ. ë‚ÂÚÓ‚Ó‰ ÔÂ‰ Ó˜ËÒÚÍÓÈ Ë/ËÎË ‰ÂÁËÌÙÂ͈ËÂÈ Ò‚ÂÚÓ‚Ó‰‡ Â„Ó ÒΉÛÂÚ Ô‰‚‡ËÚÂθÌÓ Ó·‡-·ÓÚ‡Ú¸. ùÚÓ Ô‡‚ËÎÓ ‰ÂÈÒÚ‚ÛÂÚ Ë ‚ ÒÎÛ˜‡Â χ¯ËÌÌÓÈ, Ë ‚ ÒÎÛ˜‡Â Û˜ÌÓÈ Ó˜ËÒÚÍË Ë ‰ÂÁËÌÙÂ͈ËË. àÌÚÂÌÒ˂̇fl Ó˜ËÒÚ͇ Ë ÒÚÂËÎËÁ‡ˆËfl ÌÂÓ·ıÓ‰ËÏ˚ ‰Îfl ÚÓ„Ó, ˜ÚÓ·˚ ÔÓÒÎÂ‰Û˛˘‡fl ÒÚÂ-ËÎËÁ‡ˆËfl ·˚· ˝ÙÙÂÍÚË‚ÌÓÈ. èÓʇÎÛÈÒÚ‡, ÔÓθÁÛÈÚÂÒ¸ ËÒÍβ˜ËÚÂθÌÓ ÚÓθÍÓ Ô‡Ó-‚ÓÈ ÒÚÂËÎËÁ‡ˆËÂÈ. ÇÂÏfl ÒÚÂËÎËÁ‡ˆËË (‚ÂÏfl ˝ÍÒÔÓÁˈËË ÔË ÚÂÏÔÂ‡ÚÛ ÒÚÂËÎË-Á‡ˆËË) ÒÓÒÚ‡‚ÎflÂÚ 4 ÏËÌÛÚ˚ ÔË 134°C; ‰‡‚ÎÂÌË ‰ÓÎÊÌÓ ÒÓÒÚ‡‚ÎflÚ¸ 2 ·‡‡. èÓÒÛ¯Ë-Ú ÒÚÂËÎËÁÓ‚‡ÌÌ˚È Ò‚ÂÚÓ‚Ó‰ ËÎË Ò ÔÓÏÓ˘¸˛ ÒÔˆˇθÌÓÈ ÔÓ„‡ÏÏ˚ ÔÓÒۯ˂‡ÌËfl ‚‡¯Â„Ó Ô‡Ó‚Ó„Ó ‡‚ÚÓÍ·‚‡ ËÎË Ò ÔÓÏÓ˘¸˛ „Ófl˜Â„Ó ‚ÓÁ‰Ûı‡. ë‚ÂÚÓ‚Ó‰ ÔÓ ÂÁÛÎ¸Ú‡Ú‡Ï ÚÂÒÚÓ‚ ‚˚‰ÂÊË‚‡ÂÚ ‰Ó 200 ˆËÍÎÓ‚ ÒÚÂËÎËÁ‡ˆËË. á‡ÚÂÏ ÔÓ‚Â¸Ú ҂ÂÚÓ‚Ó‰ ̇ ̇΢ˠÔÓ‚ÂʉÂÌËÈ. ÑÎfl ˝ÚÓ„Ó ÌÂÓ·ıÓ‰ËÏÓ ÔÓÒÏÓÚÂÚ¸ Â„Ó Ì‡ ÔÓÒ‚ÂÚ. ÖÒÎË ÔË ˝ÚÓÏ ‚ˉÌ˚ ÓÚ‰ÂθÌ˚ ˜ÂÌ˚ Ò„ÏÂÌÚ˚ – Á̇˜ËÚ, ˝ÚÓ ÏÂÒÚ‡ ÔÂÂÎÓÏÓ‚ ÒÚÂÍÎÓ‚ÓÎÓÍ̇, Ò‚ÂÚÓ‚Ó‰ ÒΉÛÂÚ Á‡ÏÂÌËÚ¸ ̇ ÌÓ‚˚È. äÓÌÚ‡ÍÚ˚ ‡ÍÍÛÏÛÎflÚÓ‡ ÑÎfl Ó·ÂÒÔ˜ÂÌËfl ÔÓÒÚÓflÌÌÓÈ ıÓÓ¯ÂÈ ÔÓ‚Ó‰ËÏÓÒÚË ÍÓÌÚ‡ÍÚ˚ ‡ÍÍÛÏÛÎflÚÓ‡ ÒΉÛÂÚ Á‡˘Ë˘‡Ú¸ ÓÚ Á‡„flÁÌÂÌËÈ (̇ÔËÏÂ,ÓÒÚ‡ÚÍÓ‚ ÍÓÏÔÓÁËÚ‡). ÑÎfl ˝ÚÓ„Ó ÍÓÌÚ‡ÍÚ˚ ÒΉÛÂÚ „ÛÎflÌÓ Ó˜Ë˘‡Ú¸ ‚ ‡Ï͇ı Ó·˚˜ÌÓÈ ‰ÂÁËÌÙÂ͈ËË (ÔÓÒÎÂ Í‡Ê‰Ó„Ó Ô‡ˆËÂÌÚ‡). è‰‚‡ËÚÂθ̇fl Ó·‡·ÓÚ͇ – ìÒÚ‡ÌËÚ „Û·˚ Á‡„flÁÌÂÌËfl Ò‡ÁÛ ÔÓÒΠÔËÏÂÌÂÌËfl, Ò‡ÏÓ ÔÓÁ‰Ì ˜ÂÂÁ ‰‚‡ ˜‡Ò‡ ÔÓÒΠÔËÂχ Ô‡ˆËÂÌÚ‡. í˘‡ÚÂθÌÓ ÔÓÏÓÈÚ ҂ÂÚÓ‚Ó‰ ÔÓ‰ ÔÓÚÓ˜ÌÓÈ ‚Ó‰ÓÈ (Ì ÏÂÌ 10 ÒÂÍÛ̉). åÓÊÌÓ Ú‡ÍÊ ËÒÔÓθÁÓ‚‡Ú¸ ÔÓ‰ıÓ‰fl˘Â ‰ÂÁËÌÙÂ͈ËÓÌÌÓ Ò‰ÒÚ‚Ó, Ì ÒÓ‰Âʇ˘Â ‡Î¸‰Â„ˉӂ, ˜ÚÓ·˚ ÔÂÔflÚÒÚ‚Ó‚‡Ú¸ ÔËÒ˚ı‡Ì˲ ÍÓ‚Ë. – óÚÓ·˚ Û‰‡ÎËÚ¸ Á‡„flÁÌÂÌËfl ‚Û˜ÌÛ˛, ËÒÔÓθÁÛÈÚ ÎÛ˜¯Â ‚ÒÂ„Ó Ïfl„ÍÛ˛ ˘ÂÚÍÛ ËÎË Ïfl„ÍÛ˛ Ò‡ÎÙÂÚÍÛ. èÓÎËÏÂËÁÓ‚‡ÌÌ˚È ÍÓÏÔÓÁËÚ ÏÓÊÌÓ Û‰‡ÎËÚ¸ ÒÔËÚÓÏ ËÎË Ô·ÒÚχÒÒÓ‚˚Ï ¯Ô‡ÚÂÎÂÏ. ç ËÒÔÓθÁÓ‚‡Ú¸ ÌË͇ÍËı ÓÒÚ˚ı Ô‰ÏÂÚÓ‚, ÍÓÚÓ˚ ÏÓ„ÛÚ ÔÓˆ‡‡Ô‡Ú¸ ÔÓ‚ÂıÌÓÒÚ¸. é˜ËÒÚ͇ Ë ‰ÂÁËÌÙÂ͈Ëfl ÑÎfl Ó˜ËÒÚÍË ÔÓÎÓÊËÚ ҂ÂÚÓ‚Ó‰ ‚ Ó˜Ë˘‡˛˘ËÈ ‡ÒÚ‚Ó, Ú‡Í, ˜ÚÓ·˚ ÓÌ ÔÓÎÌÓÒÚ¸˛ ·˚Î ÔÓÍ˚Ú ‡ÒÚ‚ÓÓÏ (ÛθÚ‡Á‚ÛÍÓ‚‡fl ‚‡Ì̇ ËÎË ÓÒÚÓÓÊÌÓÂ Ó˜Ë˘ÂÌË Ïfl„ÍÓÈ ˘ÂÚÍÓÈ ìÚËÎËÁ‡ˆËfl èË·Ó ÌÂθÁfl ÛÚËÎËÁËÓ‚‡Ú¸ Ò Ó·˚˜Ì˚Ï ·˚ÚÓ‚˚Ï ÏÛÒÓÓÏ. ç„ӉÌ˚ ÔË·Ó˚ ‰Îfl ÔÓÎËÏÂËÁ‡ˆËË Ë ‡ÍÍÛÏÛÎflÚÓ˚ ÒΉÛÂÚ ÛÚËÎËÁËÓ‚‡Ú¸ ‚ ÒÓÓÚ‚ÂÚÒÚ‚ËË Ò Ì‡ˆËÓ̇θÌ˚ÏË Á‡ÍÓÌÓ‰‡ÚÂθÌ˚ÏË ÌÓχÏË. ÄÍÍÛÏÛÎflÚÓ˚ ÌË ‚ ÍÓÂÏ ÒÎÛ˜‡Â Ì ·Ó-Ò‡Ú¸ ‚ Ó„Ó̸! 83 óÚÓ ‰Â·ڸ, ÂÒÎË …? ëËÏ‚ÓÎ è˘Ë̇ ìÒÚ‡ÌÂÌË ÑÂÙÂÍÚ ˝ÎÂÍÚÓÌËÍË ‚ ̇ÍÓ̘ÌËÍ ËÎË ‡ÍÍÛÏÛÎflÚÓ ÄÍÍÛÏÛÎflÚÓ ‚˚ÌÛÚ¸ Ë ÒÌÓ‚‡ ‚ÒÚ‡‚ËÚ¸. ÖÒÎË Ó¯Ë·Í‡ ÒÓı‡ÌflÂÚÒfl, Á‡ÏÂÌËÚ¸ ‡ÍÍÛÏÛÎflÚÓ ÒÂÚ‚˚Ï ÔËÚ‡ÌËÂÏ (ÙÛÌ͈Ëfl Click & Cure). ÖÒÎË Ó¯Ë·Í‡ ÒÌÓ‚‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û. èË·Ó ÔÂ„ÂÎÒfl ËÎË ÔÂÂÓı·‰ËÎÒfl чڸ ÔË·ÓÛ Óı·‰ËÚ¸Òfl (ËÎË ÔË ÔÂÂÓı·ʉÂÌÌÓÏ ÔË·Ó ‰Óʉ‡Ú¸Òfl, ÔÓ͇ ÓÌ ‰ÓÒÚË„ÌÂÚ ÍÓÏ̇ÚÌÓÈ ÚÂÏÔÂ‡ÚÛ˚) Ë ˜ÂÂÁ ÌÂÍÓÚÓÓ ‚ÂÏfl ÔÓÔ˚Ú‡Ú¸Òfl ¢ ‡Á. ÖÒÎË ÌÂÔÓ·‰Í‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û. ÑÂÙÂÍÚ ˝ÎÂÍÚÓÌËÍË ‡ÍÍÛÏÛÎflÚÓ‡ ÄÍÍÛÏÛÎflÚÓ ‚˚ÌÛÚ¸ Ë ÒÌÓ‚‡ ‚ÒÚ‡‚ËÚ¸. ÖÒÎË Ó¯Ë·Í‡ ÒÓı‡ÌflÂÚÒfl, ‚ÒÚ‡‚ËÚ¸ ̇ÍÓ̘ÌËÍ ‚ Á‡fl‰Ì˚È ·ÎÓÍ. ÖÒÎË Ó¯Ë·Í‡ ÒÓı‡ÌflÂÚÒfl, Á‡ÏÂÌËÚ¸ ‡ÍÍÛÏÛÎflÚÓ ÒÂÚ‚˚Ï ÔËÚ‡ÌËÂÏ (ÙÛÌ͈Ëfl Click & Cure). é·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û. ÄÍÍÛÏÛÎflÚÓ ‡ÁflÊÂÌ ç‡ÍÓ̘ÌËÍ ‚ÒÚ‡‚ËÚ¸ ‚ Á‡fl‰Ì˚È ·ÎÓÍ. ç‡ ‰ËÒÔΠÔÓfl‚ËÚÒfl Í‡ÚÍÓ‚ÂÏÂÌ̇fl ̇‰ÔËÒ¸ „Charging“ ÔÂʉ ˜ÂÏ ‰ËÒÔÎÂÈ ÔÂÂȉÂÚ ‚ ÂÊËÏ Stand-by. ÖÒÎË ‡ÍÍÛÏÛÎflÚÓ Ì Á‡flʇÂÚÒfl, ÒΉÛÂÚ ÔÓ˜ËÒÚËÚ¸ ÍÓÌÚ‡ÍÚ˚. ÖÒÎË ÌÂÔÓ·‰Í‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û. èË Á‡fl‰Í ̇ ‰ËÒÔΠ̠ÔÓfl‚ÎflÂÚÒfl ÌË͇ÍËı ËÁÓ·‡ÊÂÌËÈ á‡flʇ˛˘Ë ÍÓÌÚ‡ÍÚ˚ Á‡„flÁÌÂÌ˚ èÓÍÓÌÚÓÎËÓ‚‡Ú¸, Ô‡‚ËθÌÓ ÎË ‚ÒÚ‡‚ÎÂÌ Ì‡ÍÓ̘ÌËÍ ‚ Á‡fl‰Ì˚È ·ÎÓÍ. èÓ˜ËÒÚËÚ¸ ÍÓÌÚ‡ÍÚ˚ ‡ÍÍÛÏÛÎflÚÓ‡. ÑÎfl ËÌÙÓχˆËË: ç‡ ‰ËÒÔΠÔÓfl‚ËÚÒfl Í‡ÚÍÓ‚ÂÏÂÌ̇fl ̇‰ÔËÒ¸ “Charging“ ÔÂʉ ˜ÂÏ ‰ËÒÔÎÂÈ ÔÂÂȉÂÚ ‚ ÂÊËÏ Stand-by. ë‚ÂÚÓ‰ËÓ‰ Á‡fl‰ÌÓ„Ó ·ÎÓ͇ Ì „ÓËÚ ëÂÚ‚ÓÈ ·ÎÓÍ Ì ÔÓ‰Íβ˜ÂÌ ËÎË ‰ÂÙÂÍÚ ÒÂÚÂ‚Ó„Ó ·ÎÓ͇. 84 ÄÍÍÛÏÛÎflÚÓ ‚ ·ÂÁÓÔ‡ÒÌÓÏ ÂÊËÏ (‚ÒÚÓÂÌÌ˚È Á‡˘ËÚÌ˚È ‚˚Íβ˜‡ÚÂθ) ÖÒÎË Ó˜ËÒÚ͇ ÍÓÌÚ‡ÍÚÓ‚ ÛÒÔÂı‡ Ì ÔËÌÂÒ·, ‡ÍÍÛÏÛÎflÚÓ ÒΉÛÂÚ ‚˚ÌÛÚ¸. á‡ÚÂÏ ‡ÍÍÛÏÛÎflÚÓ ÓÚ‰ÂθÌÓ ÓÚ Ì‡ÍÓ̘ÌË͇ Á‡flʇڸ ‚ Á‡fl‰ÌÓÏ ·ÎÓÍ ÔËÏÂÌÓ 10 ÏËÌÛÚ. ÖÒÎË ÌÂÔÓ·‰Í‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û. èÓÍÓÌÚÓÎËÓ‚‡Ú¸, Ô‡‚ËθÌÓ ÎË ÔÓ‰ÒÓ‰ËÌÂÌ ÒÂÚ‚ÓÈ ·ÎÓÍ Í Á‡fl‰ÌÓÏÛ ·ÎÓÍÛ, ËÎË ÔÓ‰ÒÓ‰ËÌÂÌ ÎË ÒÂÚ‚ÓÈ ·ÎÓÍ Í ÒÂÚË ÒÂÚ‚˚Ï Í‡·ÂÎÂÏ (ÔË Ô‡‚ËθÌÓÈ ‡·ÓÚ ̇ ÒÂÚ‚ÓÏ ·ÎÓÍ „ÓËÚ ÁÂÎÂ̇fl ·ÏÔӘ͇). ÖÒÎË ÌÂÔÓ·‰Í‡ ÒÓı‡ÌflÂÚÒfl, Ó·‡ÚËÚÂÒ¸ Í ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û. ɇ‡ÌÚËfl / ÑÂÈÒÚ‚Ëfl ‚ ÒÎÛ˜‡Â ÂÏÓÌÚ‡ ɇ‡ÌÚËÈÌ˚È ÒÓÍ ‰Îfl ÔË·Ó‡ Bluephase N ÒÓÒÚ‡‚ÎflÂÚ 3 „Ó‰‡ ÒÓ ‰Ìfl ÔÓÍÛÔÍË (̇ ‡ÍÍÛÏÛÎflÚÓ – 1 „Ó‰). Ç ÒÎÛ˜‡Â ÌÂÔÓ·‰ÓÍ, Ó·ÛÒÎÓ‚ÎÂÌÌ˚ı ‰ÂÙÂÍÚÓÏ Ï‡ÚÂˇ· ËÎË Ó¯Ë·ÍÓÈ ÔË ËÁ„ÓÚÓ‚ÎÂÌËË, „‡‡ÌÚËfl Ó·ÂÒÔ˜˂‡ÂÚ ·ÂÒÔ·ÚÌ˚È ÂÏÓÌÚ ‡ÔÔ‡‡Ú‡. ë‚Âı ˝ÚÓ„Ó „‡‡ÌÚËfl Ì ‰‡ÂÚ Ô‡‚‡ ̇ ‚ÓÁÏ¢ÂÌË χÚÂˇθÌÓ„Ó ËÎË ÏÓ‡Î¸ÌÓ„Ó Û˘Â·‡. èË ˝ÚÓÏ ÔË·Ó ‰ÓÎÊÂÌ ËÒÔÓθÁÓ‚‡Ú¸Òfl ËÒÍβ˜ËÚÂθÌÓ ÚÓθÍÓ ÔÓ Ì‡Á̇˜ÂÌ˲. ã˛·Ó ËÒÔÓθÁÓ‚‡ÌË ‚ ËÌ˚ı ˆÂÎflı fl‚ÎflÂÚÒfl ËÒÔÓθÁÓ‚‡ÌËÂÏ Ì ÔÓ Ì‡Á̇˜ÂÌ˲ – Á‡ ÔÓÎÛ˜ÂÌÌ˚ ÂÁÛθڇÚ˚ ÔÓËÁ‚Ó‰ËÚÂθ Ì ÌÂÒÂÚ ÓÚ‚ÂÚÒÚ‚ÂÌÌÓÒÚË Ë Ì ӷÂÒÔ˜˂‡ÂÚ „‡‡ÌÚ˲. ä Ú‡ÍËÏ ÒÎÛ˜‡flÏ ÓÚÌÓÒflÚÒfl: – Û˘Â·, ̇ÌÂÒÂÌÌ˚È ‚ ÂÁÛθڇÚ ÌÂÍÓÂÍÚÌÓ„Ó Ó·‡˘ÂÌËfl Ò ÔË·ÓÓÏ. éÒÓ·ÂÌÌÓ ˝ÚÓ ÓÚÌÓÒËÚÒfl Í ÌÂÔ‡‚ËθÌÓÏÛ ı‡ÌÂÌ˲ ‡ÍÍÛÏÛÎflÚÓÓ‚ (ÒÏ. íÂıÌ˘ÂÒÍË ‰‡ÌÌ˚Â: ìÒÎÓ‚Ëfl Ú‡ÌÒÔÓÚËÓ‚ÍË Ë ı‡ÌÂÌËfl). – ÔÓ‚ÂʉÂÌË ‰ÂÚ‡ÎÂÈ, ÍÓÚÓ˚ ÔÓ‰ÎÂÊ‡Ú ËÁÌÓÒÛ ‚ ÂÁÛθڇÚ ÌÓχθÌÓÈ ‡·ÓÚ˚ (̇ÔËÏÂ, ‡ÍÍÛÏÛÎflÚÓ). – ÔÓ‚ÂʉÂÌËfl ‚ ÂÁÛθڇÚ ‚̯ÌËı ‚ÓÁ‰ÂÈÒÚ‚ËÈ, ̇ÔËÏÂ, Û‰‡‡, Ô‡‰ÂÌËfl ̇ ÔÓÎ – ÔÓ‚ÂʉÂÌËfl ËÁ-Á‡ ÌÂÍÓÂÍÚÌÓÈ ÛÒÚ‡ÌÓ‚ÍË ÎË·Ó ËÌÒÚ‡ÎÎflˆËË ÔË·Ó‡ – ÔÓ‚ÂʉÂÌËfl, ÔÓÎÛ˜ÂÌÌ˚ ‚ ÂÁÛθڇÚ ÔÓ‰Íβ˜ÂÌËfl ÔË·Ó‡ Í ÒÂÚË Ò Ì‡ÔflÊÂÌËÂÏ Ë ˜‡ÒÚÓÚÓÈ, ÓÚ΢Ì˚ÏË ÓÚ Û͇Á‡ÌÌ˚ı ̇ Ú‡·Î˘Í ÔË·Ó‡. – ÔÓ‚ÂʉÂÌËfl, ÔÓÎÛ˜ÂÌÌ˚ ‚ ÂÁÛθڇÚ ÂÏÓÌÚÌ˚ı ‡·ÓÚ ÎË·Ó ËÁÏÂÌÂÌËÈ ‚ ÔË·ÓÂ, ÍÓÚÓ˚ ·˚ÎË ÔÓËÁ‚‰ÂÌ˚ Ó„‡ÌËÁ‡ˆËflÏË, Ì Ëϲ˘ËÏË ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘ÂÈ ‡‚ÚÓËÁ‡ˆËË. ÖÒÎË ÒÎÛ˜‡È ÔËÁÌ‡Ì „‡‡ÌÚËÈÌ˚Ï, ‚ÂÒ¸ ÔË·Ó ‚ Ò·Ó (̇ÍÓ̘ÌËÍ, Á‡fl‰Ì˚È ·ÎÓÍ, ‡ÍÍÛÏÛÎflÚÓ, ÒÂÚ‚ÓÈ Í‡·Âθ, ÒÂÚ‚ÓÈ ·ÎÓÍ) ÒΉÛÂÚ ÔÓÒ·ڸ ‚ÏÂÒÚÂ Ò ‰ÓÍÛÏÂÌÚÓÏ Ó· ÓÔ·Ú ÔË·Ó‡ ‚ ÓË„Ë̇θÌÓÈ ÛÔ‡ÍÓ‚ÍÂ Ò ÒÓÓÚ‚ÂÚÒÚ‚Û˛˘ËÏË Í‡ÚÓÌÌ˚ÏË ‚Í·‰Í‡ÏË (‰ÓÒÚ‡‚ÍÛ ÓÔ·˜Ë‚‡ÂÚ ÔÓÚ·ËÚÂθ) ÔÓ‰‡‚ˆÛ ËÎË Ì‡ÔflÏÛ˛ ̇ Vivadent Ivoclar. ã˛·˚ ÂÏÓÌÚÌ˚ ‡·ÓÚ˚ ÏÓ„ÛÚ ÔÓËÁ‚Ó‰ËÚ¸Òfl ÚÓθÍÓ Í‚‡ÎËÙˈËÓ‚‡ÌÌ˚Ï ÔÂÒÓ̇ÎÓÏ ÒÂ‚ËÒÌÓÈ ÒÎÛÊ·˚, Ëϲ˘ÂÈ ‡‚ÚÓËÁ‡ˆË˛ ÓÚ Ivoclar Vivadent. Ç ÒÎÛ˜‡Â ‚ÓÁÌËÍÌÓ‚ÂÌËfl ‰ÂÙÂÍÚ‡, ÍÓÚÓ˚È Ì ÏÓÊÂÚ ·˚Ú¸ ÛÒÚ‡ÌÂÌ Ç‡ÏË, Ó·‡ÚËÚÂÒ¸, ÔÓʇÎÛÈÒÚ‡, Í Ç‡¯ÂÏÛ ÔÓ‰‡‚ˆÛ ËÎË ‚ ÒÂ‚ËÒÌÛ˛ ÒÎÛÊ·Û (‡‰ÂÒ‡ Ç˚ ̇ȉÂÚ ̇ Ó·ÎÓÊÍ ËÌÒÚÛ͈ËË). óÂÚÍÓ ÓÔËÒ‡ÌË ‰ÂÙÂÍÚ‡ ËÎË Ó·ÒÚÓflÚÂθÒÚ‚, ÍÓÚÓ˚ ÔË‚ÂÎË Í ‰ÂÙÂÍÚÛ, ӷ΄˜‡˛Ú ÔÓËÒÍ ÌÂÔÓ·‰ÍË. èÓʇÎÛÈÒÚ‡, ÔËÎÓÊËÚ ˝ÚÓ ÓÔËÒ‡ÌËÂ Í ‡ÔÔ‡‡ÚÛ. ëÔˆËÙË͇ˆËfl ëÔˆËÙË͇ˆËfl 1 Á‡fl‰Ì˚È ·ÎÓÍ, 1 ÒÂÚ‚ÓÈ Í‡·Âθ, 1 ÒÂÚ‚ÓÈ ·ÎÓÍ, 1 ̇ÍÓ̘ÌËÍ, 1 ‡ÍÍÛÏÛÎflÚÓ (Li-Po), 1 Ò‚ÂÚÓ‚Ó‰ 10 ÏÏ, 3 Á‡˘ËÚÌ˚ı ˝Í‡Ì‡, 1 ËÌÒÚÛ͈Ëfl èË̇‰ÎÂÊÌÓÒÚË ÑÎfl LJ¯ÂÈ Î‡ÏÔ˚ Bluephase N ‚˚ÔÛÒ͇˛ÚÒfl ÒÎÂ‰Û˛˘Ë ÔË̇‰ÎÂÊÌÓÒÚË: ÄÚ.‹ ç‡ËÏÂÌÓ‚‡ÌË 656196 ë‚ÂÚÓ‚Ó‰ 10 ÏÏ ˜ÂÌ˚È (N) 608538 ë‚ÂÚÓ‚Ó‰ Pin-point 6>2 ÏÏ ˜ÂÌ˚È 551756 ᇢËÚÌ˚È ÍÓÁ˚ÂÍ (ÔÂ‚ÂÌÛÚ˚È ÍÓÌÛÒ) 592496 ᇢËÚÌ˚È ˝Í‡Ì 659956 ÄÍÍÛÏÛÎflÚÓ Bluephase N 659681 ç‡ÍÓ̘ÌËÍ Bluephase N 607922 Bluephase Meter 608554 Bluephase Á‡˘ËÚÌ˚ ˜ÂıÎ˚ íÂıÌ˘ÂÒÍË ‰‡ÌÌ˚ ꇷӘ ̇ÔflÊÂÌË Á‡fl‰ÌÓ„Ó ·ÎÓ͇ ꇷӘ ̇ÔflÊÂÌË ̇ÍÓ̘ÌË͇ ëÂÚ‚ÓÈ ·ÎÓÍ ç‡ ‚˚ıӉ íËÔ 15.2630 5 VDC 3.7 VDC Ò ‡ÍÍÛÏÛÎflÚÓÓÏ 5 VDC Ò ÒÂÚ‚˚Ï ·ÎÓÍÓÏ 100–240 VAC / 50-60 Ɉ / χÍÒ. 0.4 A / ÔÓËÁ‚Ó‰ËÚÂθ: Friwo 5 VDC ìÒÎÓ‚Ëfl ˝ÍÒÔÎÛ‡Ú‡ˆËË íÂÏÔÂ‡ÚÛ‡ éÚÌÓÒËÚÂθ̇fl ‚·ÊÌÓÒÚ¸ ч‚ÎÂÌË ê‡ÁÏÂ˚ Á‡fl‰ÌÓ„Ó ·ÎÓ͇ ÇÂÒ Á‡fl‰ÌÓ„Ó ·ÎÓ͇ ÇÂÏfl Á‡fl‰ÍË ‡ÍÍÛÏÛÎflÚÓ‡ ÓÚ +10 °C ‰Ó +35 °C ÓÚ 30% ‰Ó 75% ÓÚ 700 hPa ‰Ó 1060 hPa Ñ=205 ÏÏ; ò=150 ÏÏ; Ç=85 ÏÏ 250 „ ÔËÏ. 2 ˜ (ÔË ÔÓÎÌÓÒÚ¸˛ ‡ÁflÊÂÌÌÓÏ ‡ÍÍÛÏÛÎflÚÓÂ) ùÎÂÍÚÓÔËÚ‡ÌËfl ̇ÍÓ̘ÌË͇ ‡ÍÍÛÏÛÎflÚÓ Li-Po å‡ÍÒ. ‡·Ó˜Â ‚ÂÏfl ‡ÍÍÛÏÛÎflÚÓ‡ ÔËÏ. 60 ÏËÌ. (ÔË ÌÓ‚ÓÏ, ÔÓÎÌÓÒÚ¸˛ Á‡flÊÂÌÌÓÏ ‡ÍÍÛÏÛÎflÚÓÂ) àÒÚÓ˜ÌËÍ Ò‚ÂÚ‡ ÔÓÎË‚ÓÎÌÓ‚ÓÈ ‰ËÓ‰ ÑˇԇÁÓÌ Ò‚ÂÚÓ‚ÓÈ ‚ÓÎÌ˚ 385–515 ÌÏ åÓ˘ÌÓÒÚ¸ Ò‚ÂÚÓ‚Ó„Ó ËÁÎÛ˜ÂÌËfl χÍÒ. 1.200 ÏÇÚ/ÒÏ2 ± 10% êÂÊËÏ ‡·ÓÚ˚ 5 ÏËÌ. ‚Íβ˜ÂÌ / 6 ÏËÌ. ‚˚Íβ˜ÂÌ (Ò Ô‡ÛÁ‡ÏË) ê‡ÁÏÂ˚ ̇ÍÓ̘ÌË͇: Ñ=260 ÏÏ; ò=42 ÏÏ; Ç=120 ÏÏ ÇÂÒ Ì‡ÍÓ̘ÌË͇ (c ‡ÍÍÛÏÛÎflÚÓÓÏ) 225 „ ìÒÎÓ‚Ëfl Ú‡ÌÒÔÓÚËÓ‚ÍË Ë ı‡ÌÂÌËfl: íÂÏÔÂ‡ÚÛ‡ ÓÚ -20 °C ‰Ó +60 °C éÚÌÓÒËÚÂθ̇fl ‚·ÊÌÓÒÚ¸ ÓÚ 10% ‰Ó 75% ч‚ÎÂÌË ÓÚ 500 hPa ‰Ó 1060 hPa ã‡ÏÔÛ Bluephase N ı‡ÌËÚ¸ ‚ Á‡Í˚ÚÓÏ ÔÓÏ¢ÂÌËË. èË·Ó Ì ÔÓ‰‚Â„‡Ú¸ ÒËθÌ˚Ï ÒÓÚflÒÂÌËflÏ ÄÍÍÛÏÛÎflÚÓ – Ì ı‡ÌËÚ¸ ÔË ÚÂÏÔÂ‡ÚÛ ‚˚¯Â 40°C (ËÎË Í‡ÚÍÓ‚ÂÏÂÌÌÓ ÔË 60°C), ÂÍÓÏẨÛÂχfl ÚÂÏÔÂ‡ÚÛ‡ ı‡ÌÂÌËfl 15 – 30°C. – ı‡ÌËÚ¸ ‚Ò„‰‡ Á‡flÊÂÌÌ˚Ï Ë – Ì ‰Óθ¯Â 6 ÏÂÒflˆÂ‚ 85 Appendix Guidance and manufacturer‘s declaration - electromagnetic emission The “Bluephase N” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase N” should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment - guidance RF emissions CISPR 11 Group 1 The “Bluephase N” uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emissions CISPR 11 Class B The “Bluephase N” is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic emissions IEC 61000-3-2 Class A P < 75W (pass without test) Voltage fluctuations / flicker emissions IEC 61000-3-3 Complies P < 75W (pass without test) Guidance and manufacturer‘s declaration - electromagnetic immunity The “Bluephase N” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase N” should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Electrostatic discharge (ESD) IEC 61000-4-2 ± 6 kV contact ± 8 kV air ± 6 kV contact ± 8 kV air Floors should be concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Electrical fast transient/burst IEC 61000-4-4 ± 2 kV for power supply lines ± 1 kV for input/output lines ± 2 kV for power supply lines Mains power quality should be that of typical commercial or dental environment. ± 1 kV for input/output lines Surge IEC 61000-4-5 ± 1 kV line - line ± 2 kV line - earth ± 1 kV line - line no prot. earth Mains power quality should be that of typical commercial or dental environment. Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 <5 % U (>95 % dip in U ) for 0.5 cycle <5 % U (>95 % dip in U ) for 0.5 cycle 40 % U (60 % dip in U ) for 5 cycles 40 % U (60 % dip in U ) for 5 cycles Mains power quality should be that of typical commercial or dental environment. If the user of the „Bluephase N“ requires continued operation during power mains interruptions, it is recommended that the „Bluephase N“ be powered from an uninterruptible power supply or battery. 70 % U (30 % dip in U ) for 25 cycles 70 % U (30 % dip in U ) for 25 cycles <5 % U (>95 % dip in U ) for 5 sec <5 % U (>95 % dip in U ) for 5 sec 3 A/m 30 A/m T T T T T T T T Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 T T T T T T T T NOTE: UT is the a.c. mains voltage prior to application of the test level. 86 Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or dental environment. Guidance and manufacturer‘s declaration - electromagnetic immunity The “Bluephase N” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase N” should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should not be used closer to any part of the „Bluephase N“, including cables, than the recommended separation distance caculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 3 Vrms 150 kHz to 80 MHz 10 V d = 0.35 P Radiated RF IEC 61000-4-3 3 V/m 80 MHz to 2.5 GHz 10 V/m d = 0.35 P 80 MHz to 800 MHz d = 0.70P 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the „Bluephase N“ is used exceeds the applicable RF compliance level above, the „Bluephase N“ should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the „Bluephase N“. b Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m. Recommended separation distances between portable and mobile RF communications equipment and the “Bluephase N” The “Bluephase N” is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the “Bluephase N” can help prevent electromagnetic interference by maintaining a miminum distance between portable and mobile RF communications equipment (transmitters) and the „Bluephase N“ as recommended below, according to the maximum output power of the communication equipment. Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d = 0.4 P 80 MHz to 800 MHz d = 0.4 P 800 MHz to 2.5 GHz d = 0.7 P 0.01 0.04 0.04 0.07 0.1 0.13 0.13 0.22 1 0.40 0.40 0.70 10 1.3 1.3 2.2 100 4.0 4.0 7.0 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: A  t 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 4: T hese guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. 87 Ivoclar ­Vivadent – worldwide Ivoclar Vivadent AG Bendererstrasse 2 9494 Schaan Liechtenstein Tel. +423 235 35 35 Fax +423 235 33 60 www.ivoclarvivadent.com Ivoclar Vivadent SAS B.P. 118 F-74410 Saint-Jorioz France Tel. +33 4 50 88 64 00 Fax +33 4 50 68 91 52 www.ivoclarvivadent.fr Ivoclar Vivadent Pty. Ltd. 1 – 5 Overseas Drive P.O. Box 367 Noble Park, Vic. 3174 Australia Tel. +61 3 9795 9599 Fax +61 3 9795 9645 www.ivoclarvivadent.com.au Ivoclar Vivadent GmbH Dr. Adolf-Schneider-Str. 2 D-73479 Ellwangen, Jagst Germany Tel. +49 7961 889 0 Fax +49 7961 6326 www.ivoclarvivadent.de Ivoclar Vivadent GmbH Tech Gate Vienna Donau-City-Strasse 1 1220 Wien Austria Tel. +43 1 263 191 10 Fax: +43 1 263 191 111 www.ivoclarvivadent.at Ivoclar Vivadent Ltda. Alameda Caiapós, 723 Centro Empresarial Tamboré CEP 06460-110 Barueri – SP Brazil Tel. +55 11 2424 7400 Fax +55 11 3466 0840 www.ivoclarvivadent.com.br Ivoclar Vivadent Inc. 1-6600 Dixie Road Mississauga, Ontario L5T 2Y2 Canada Tel. +1 905 670 8499 Fax +1 905 670 3102 www.ivoclarvivadent.us Ivoclar Vivadent Shanghai Trading Co., Ltd. 2/F Building 1, 881 Wuding Road, Jing An District 200040 Shanghai China Tel. +86 21 6032 1657 Fax +86 21 6176 0968 www.ivoclarvivadent.com Ivoclar Vivadent Marketing Ltd. Calle 134 No. 7-B-83, Of. 520 Bogotá Colombia Tel. +57 1 627 3399 Fax +57 1 633 1663 www.ivoclarvivadent.co Wieland Dental + Technik GmbH & Co. KG Lindenstrasse 2 75175 Pforzheim Germany Tel. +49 7231 3705 0 Fax +49 7231 3579 59 www.wieland-dental.com Ivoclar Vivadent Marketing (India) Pvt. Ltd. 503/504 Raheja Plaza 15 B Shah Industrial Estate Veera Desai Road, Andheri (West) Mumbai, 400 053 India Tel. +91 22 2673 0302 Fax +91 22 2673 0301 www.ivoclarvivadent.in Ivoclar Vivadent s.r.l. Via Isonzo 67/69 40033 Casalecchio di Reno (BO) Italy Tel. +39 051 6113555 Fax +39 051 6113565 www.ivoclarvivadent.it Ivoclar Vivadent K.K. 1-28-24-4F Hongo Bunkyo-ku Tokyo 113-0033 Japan Tel. +81 3 6903 3535 Fax +81 3 5844 3657 www.ivoclarvivadent.jp Ivoclar Vivadent Ltd. 12F W-Tower 54 Seocho-daero 77-gil, Seocho-gu Seoul, 06611 Republic of Korea Tel. +82 2 536 0714 Fax +82 2 596 0155 www.ivoclarvivadent.co.kr Date information prepared: 2016-06 / Rev. 2 657717/2016-06-09 Ivoclar Vivadent S.A. de C.V. Calzada de Tlalpan 564, Col Moderna, Del Benito Juárez 03810 México, D.F. México Tel. +52 55 50 62 10 00 Fax +52 55 50 62 10 29 www.ivoclarvivadent.com.mx Ivoclar Vivadent S.L.U. Carretera de Fuencarral nº24 Portal 1 – Planta Baja 28108-Alcobendas (Madrid) Spain Telf. +34 91 375 78 20 Fax: +34 91 375 78 38 www.ivoclarvivadent.es Ivoclar Vivadent BV De Fruittuinen 32 2132 NZ Hoofddorp Netherlands Tel. +31 23 529 3791 Fax +31 23 555 4504 www.ivoclarvivadent.com Ivoclar Vivadent AB Dalvägen 14 S-169 56 Solna Sweden Tel. +46 8 514 939 30 Fax +46 8 514 939 40 www.ivoclarvivadent.se Ivoclar Vivadent Ltd. 12 Omega St, Rosedale PO Box 303011 North Harbour Auckland 0751 New Zealand Tel. +64 9 914 9999 Fax +64 9 914 9990 www.ivoclarvivadent.co.nz Ivoclar Vivadent Liaison Office : Tesvikiye Mahallesi Sakayik Sokak Nisantas’ Plaza No:38/2 Kat:5 Daire:24 34021 Sisli – Istanbul Turkey Tel. +90 212 343 0802 Fax +90 212 343 0842 www.ivoclarvivadent.com Ivoclar Vivadent Polska Sp. z o.o. Al. Jana Pawla II 78 00-175 Warszawa Poland Tel. +48 22 635 5496 Fax +48 22 635 5469 www.ivoclarvivadent.pl Ivoclar Vivadent Marketing Ltd. Prospekt Andropova 18 korp. 6/ office 10-06 115432 Moscow Russia Tel. +7 499 418 0300 Fax +7 499 418 0310 www.ivoclarvivadent.ru Ivoclar Vivadent Marketing Ltd. Qlaya Main St. Siricon Building No.14, 2nd Floor Office No. 204 P.O. Box 300146 Riyadh 11372 Saudi Arabia Tel. +966 11 293 8345 Fax +966 11 293 8344 www.ivoclarvivadent.com Ivoclar Vivadent Limited Ground Floor Compass Building Feldspar Close Warrens Business Park Enderby Leicester LE19 4SE United Kingdom Tel. +44 116 284 7880 Fax +44 116 284 7881 www.ivoclarvivadent.co.uk Ivoclar Vivadent, Inc. 175 Pineview Drive Amherst, N.Y. 14228 USA Tel. +1 800 533 6825 Fax +1 716 691 2285 www.ivoclarvivadent.us
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88

Ivoclar Vivadent Bluephase N Instructions For Use Manual

Tipo
Instructions For Use Manual