Ohaus ST270 Manual de usuario

Tipo
Manual de usuario
General instructions for
ST270 Spear pH electrode
Instrucciones generales para el
electrodo de pH con punta de lanza ST270
Instructions gérales pour
ST270 électrode pH à pointe effilée
Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054 USA
Tel: (973) 377-9000
Fax: (973) 944-7177
With offices worldwide
Con oficinas en todo el mundo
Avec des bureaux dans le monde entier
www.ohaus.com; [email protected]
*30240979*
P/N 30240979 A © 2015,
Ohaus Corporation, all rights reserved/ todos
los derechos reservados/ tous droits réservés
EN-2 ST270 pH electrode
ST270 pH electrode EN-1
Model
Item NO
Description
ST270
30240974
2-in-1 glass shaft spear pH
electrode
Specifications:
Connection: BNC, Cinch
Cable length: 1.0m
Shaft length: 120mm
Shaft diameter: 12mm
pH range: 0-14pH
Zero potential: 7.00±0.25pH
Slope: 97%
Impedance: 100M
Shaft material: Glass
Reference Electrolyte: 3M KCl Solution
Temperature range: 0-100
Temperature probe: NTC 30kΩ
Reference Diaphragm: Sleeve
OHAUS offer different pH electrodes for
different applications. Make sure you have
chosen the best fit pH electrode for your
samples; and start by performing a calibration
of the electrode.
ST270 is a 2-in-1 pH glass electrode. A 3-in-1
electrode also has a temperature probe in it.
EN-2 ST270 pH electrode
Before using the electrode in calibration or
measurement, please note:
1. When unpacking, check the electrode for
any mechanical damage to the membrane,
diaphragm or body.
2. Screw off watering cap first, then the lid,
rinse electrode with pure water. Wipe
electrode with a tissue. Do not rub the glass
membrane, or the response time might
increase.
3. Eliminate any air bubbles inside the
membrane. Make the electrode inside
pressure in equilibrium with the atmosphere
by opening the electrode refilling aperture
during measurement.
FAQs:
1. Whats the meaning of electrode
condition icon?
After calibration, the pH meter will judge the
electrode slope and offset, then display the
judgment on the screen:
slope > 95%
Electrode condition is very good
slope : 90%-95%
Electrode condition is acceptable
slope : 85%-90%
Electrode condition is not good, need re-
calibration or replacing.
ST270 pH electrode EN-3
2. How do you properly store the pH
electrode after every use?
A: Rinse the electrode with pure water, put
the electrode into the watering cap. If the
solution in the cap is polluted or is dried,
clean it and fill it with 3M KCl solution. Avoid
placing the electrode a dry place or deionized
water for a long time.
3. Whats the life of the pH electrodes?
A: Measured samples, storage conditions
and maintenance methods all could influence
the life of pH electrodes; normally a pH
electrode expires after 12 months. Some
sample, such as strong acid or alkali,
corrosive liquid will shorten the pH electrode
life. The pH electrode is consumable and is
generally provided with warranty.
4. The key operation points for ST270?
A: During use, ST270 electrode tip generally
does not break. Because of the distance
between tip and junction is longer than
normal, the depth must exceed this distance,
to ensure valid test data.
Due to the particularity of samples, ST270
electrode must be cleaned from time to time.
For example, when measuring protein
medium, we should use the pepsin/HCl
cleaning fluid to rinse the electrode. When
measuring Oily/organic solution, electrode
should be rinsed by acetone or ethanol.
EN-4 ST270 pH electrode
Troubleshooting
If the electrode readings are abnormal, the
following procedures should be paid attention
to.
1. If the slope is below 95% it needs to be re-
calibrated.
2. If the measuring depth of electrode head in
the sample is low the junction must be in
contact with sample.
3. Soak and rinse the electrode in a cleaning
solution for 2 hours after use.
Electrodo de pH ST270 ES-1
Modelo
de
ítem
Descripción
ST270
30240974
Electrodo de pH de vidrio con
punta de lanza 2 en 1
Especificaciones:
Conexión: BNC, Cinch
Longitud del cable: 1,0 m
Longitud del astil: 120 mm
Diámetro del astil: 12 mm
Rango de pH: 0 a 14 pH
Potencial cero: 7,00 ± 0,25 pH
Pendiente: ≥ 97%
Impedancia: ≤ 100 MΩ
Material del astil: Vidrio
Electrolito de referencia: Solución 3M KCl
Rango de temperatura: 0 a 100
Sonda de temperatura: NTC 30 kΩ
Diafragma de referencia: Manga
OHAUS ofrece diferentes electrodos de pH
para diferentes aplicaciones. Asegúrese de
que ha elegido la opcn de electrodo de pH
que mejor se adapta a sus muestras, y
comience realizando una calibración del
electrodo.
El electrodo ST270 es un electrodo de pH de
vidrio 2 en1. Un electrodo 3 en 1 tiene
también una sonda de temperatura.
ES-2 Electrodo de pH ST270
Antes de utilizar el electrodo en la
calibración o en la medición, tenga en
cuenta:
1. Al desembalar el equipo, revise el
electrodo para ver si tiene algún daño
mecánico en la membrana, en el diafragma o
en el cuerpo.
2. Desatornille primero la cubierta de
hidratación, luego la tapa, enjuague el
electrodo con agua pura. Limpie el electrodo
con papel tisú. No frote la membrana de
vidrio, ya que el tiempo de respuesta podría
aumentar.
3. Elimine las burbujas de aire dentro de la
membrana. Equilibre la presión interior del
electrodo con la presión atmosférica abriendo
la abertura de relleno del electrodo durante la
medición.
Preguntas frecuentes:
1. ¿Cuál es el significado del icono de
condición del electrodo?
Después de la calibracn, el medidor de pH
estima la pendiente y la compensación del
electrodo, a continuación, luego muestra la
estimación en la pantalla:
Pendiente > 95%
La condición del electrodo es muy buena
Pendiente: 90% a 95%
La condición del electrodo es aceptable
Pendiente: 85% a 90%
La condición del electrodo no es buena, es
necesario volver a calibrarlo o sustituirlo.
Electrodo de pH ST270 ES-3
2. ¿mo se guarda correctamente el
electrodo de pH después de cada uso?
A: Enjuague el electrodo con agua pura,
ponga el electrodo dentro de la cubierta de
hidratación. Si la solución en la cubierta está
contaminada o está seca, límpiela y llénela
con solución 3M KCl. Evite colocar el
electrodo en un lugar seco o en agua
desionizada durante un largo tiempo.
3. ¿Cuál es la vida útil de los electrodos
de pH?
A: Las muestras de mediciones, las
condiciones de almacenamiento y los
todos de mantenimiento pueden afectar la
vida útil de los electrodos de pH;
normalmente, la duración de un electrodo de
pH expira es de 12 meses. Algunas
muestras, tales como ácidos o álcalis fuertes
o líquidos corrosivos acortan la vida útil del
electrodo de pH. Un electrodo de pH es un
artículo desechable y en general se
proporciona con garantía.
4. ¿Cuáles son los puntos clave de
operación del ST270?
A: Durante el uso, la punta del electrodo
ST270 en general no se rompe. Debido a
que la distancia entre la punta y la juntura es
mayor de lo normal, la profundidad debe
superar esta distancia, para garantizar la
validez de los datos.
Debido a la particularidad de las muestras, el
electrodo ST270 debe limpiarse de vez en
cuando. Por ejemplo, cuando se mide un
medio proteico, debe utilizarse un líquido de
limpieza a base de pepsina/HCl para limpiar
el electrodo. Cuando se mide una solución
aceitosa/orgánica, el electrodo debe lavarse
con acetona o etanol.
ES-4 Electrodo de pH ST270
Solución de problemas
Si las lecturas del electrodo son anormales,
deben seguirse los siguientes
procedimientos.
1. Si la pendiente es menor a 95%, es
necesario volver a calibrar.
2. Si la profundidad de medición de la
cabeza del electrodo es baja, la juntura debe
estar en contacto con la muestra.
3. Remoje y enjuague el electrodo en una
solución de limpieza durante 2 horas
después del uso.
ST270 pH electrode FR-1
Modèle
Numéro
d'article
Description
ST270
30240974
Verre d'électrode pH à pointe
effilée 2-en-1
Spécifications:
Connexion : BNC, Cinch
Longueur de câble : 1.0m
La longueur de tiges : 120mm
Diamètre de tiges: 12mm
Plage de pH : 0-14pH
Potentiel zéro : 7,00±0,25 pH
Pente : ≥ 97%
Impédance : ≤ 100MΩ
Le matériau des tiges: Le verre
Electrolyte de rérence: Solution de KCl 3 M
Plage de température: 0-100
Sonde de température: NTC 30kΩ
Diaphragme de rérence: Manchon
OHAUS offre électrodes différentes pH pour
les applications différentes. Assurez-vous
que vous avez choisi le meilleur ajustement
électrode pH pour vos échantillons; et
commencez par réaliser un étalonnage de
l'électrode.
ST270 est une électrode pH 2-en-1 en verre.
Une électrode de 3-en-1 dispose également
d'une sonde de temrature.
FR-2 ST270 électrode pH
Avant d'utiliser l'électrode pendant
l'étalonnage ou la mesure, veuillez noter :
1. Lors du déballage, vérifiez l'électrode pour
tecter tout dommage mécanique de la
membrane, le diaphragme ou de l'organisme.
2. Dévisser le capuchon d'arrosage en
premier, puis le couvercle, rincez l'électrode
avec de l'eau pure. Essuyez l'électrode avec
un mouchoir. Ne frottez pas la membrane en
verre, ou le temps de réponse pourrait
augmenter.
3. Eliminez les bulles d'air à l'intérieur de la
membrane. Faîtes l'électrode à l'intérieur à
l'équilibre de pression avec l'atmosphère en
ouvrant l'électrode remplissant l'aperture au
cours de la mesure.
Questions fréquemment posées :
1. Quelle est la signification de l'icône de
l'état de l'électrode?
Après l'étalonnage, l'oxymètre pH jugera
l'inclinaison de l'électrode et le décalage, puis
affichera l'arrêt sur l'écran :
pente > 95 %
La condition de sonde est très bonne
pente : 90%-95%
L'électrode est dans une condition
acceptable.
pente : 85%-90%
La condition de l'électrode n'est pas bonne, il
y a besoin d’un étalonnage nouveau ou du
remplacement.
ST270 pH electrode FR-3
2. Comment pouvez-vous stocker
correctement l'électrode pH aps chaque
utilisation ?
A: Rincez l'électrode avec de l'eau pure,
placez l'électrode dans le capuchon de
mouillage. Si la solution dans le capuchon est
polluée ou séchée, nettoyez-le et remplissez-
le avec 3M solution de KCl. Évitez de placer
l'électrode en un endroit sec ou dans l'eau
ionisée pour un long moment.
3. Quelle est la durée de vie des
électrodes pH?
A: Des échantillons mesurés, des méthodes
d'entretien et des conditions d'entreposage
tous pourraient influencer la vie des
électrodes pH; normalement, une électrode
pH expire après 12 mois. Certains exemples,
tels que l'acide fort ou l'alcalin liquide corrosif
raccourciront la durée de vie d'une électrode
pH. L'électrode pH est consommable et est
ralement fournie avec la garantie.
4. Les points clés de l'opération pour
ST270 ?
A: Au cours de l'utilisation, type de l'électrode
ST270 géralement ne se rompe pas. Étant
que la distance entre la pointe et la jonction
est plus longue que la normale, la profondeur
doit dépasser cette distance, afin de garantir
la validité des données du test.
En raison de la particularité des échantillons,
l'électrode ST270 doit être nettoe de
temps en temps. Par exemple, perdant la
mesure de la protéine moyenne, nous
devrions utiliser le fluide de nettoyage
pepsine/HCl pour rincer l'électrode. Lors de la
mesure de la solution organique/huileuse
l'électrode doit être rincée avec l'acétone ou
l'éthanol.
FR-4 ST270 électrode pH
pannage
Si les relevés de l'électrode sont anormaux,
vous devez prêter l'attention aux procédures
suivantes.
1. Si la pente est inférieure de 95 %, elle a
besoin d'être rétalonnée.
2. Si la mesure de la profondeur de la tête
d'électrode dans l'échantillon est faible la
jonction doit être en contact avec
l'échantillon.
3. Tremper et rincer l'électrode dans une
solution de nettoyage pendant 2 heures
après l'utilisation.

Transcripción de documentos

General instructions for ST270 Spear pH electrode Instrucciones generales para el electrodo de pH con punta de lanza ST270 Instructions générales pour ST270 électrode pH àpointe effilée Ohaus Corporation 7 Campus Drive Suite 310 Parsippany, NJ 07054 USA Tel: (973) 377-9000 Fax: (973) 944-7177 With offices worldwide Con oficinas en todo el mundo Avec des bureaux dans le monde entier www.ohaus.com; [email protected] *30240979* P/N 30240979 A © 2015, Ohaus Corporation, all rights reserved/ todos los derechos reservados/ tous droits réservés EN-2 ST270 pH electrode ST270 pH electrode Model ST270 Item NO 30240974 EN-1 Description 2-in-1 glass electrode shaft spear pH Specifications: Connection: BNC, Cinch Cable length: 1.0m Shaft length: 120mm Shaft diameter: 12mm pH range: 0-14pH Zero potential: 7.00±0.25pH Slope: ≥ 97% Impedance: ≤ 100MΩ Shaft material: Glass Reference Electrolyte: 3M KCl Solution Temperature range: 0-100℃ Temperature probe: NTC 30kΩ Reference Diaphragm: Sleeve OHAUS offer different pH electrodes for different applications. Make sure you have chosen the best fit pH electrode for your samples; and start by performing a calibration of the electrode. ST270 is a 2-in-1 pH glass electrode. A 3-in-1 electrode also has a temperature probe in it. EN-2 ST270 pH electrode Before using the electrode in calibration or measurement, please note: 1. When unpacking, check the electrode for any mechanical damage to the membrane, diaphragm or body. 2. Screw off watering cap first, then the lid, rinse electrode with pure water. Wipe electrode with a tissue. Do not rub the glass membrane, or the response time might increase. 3. Eliminate any air bubbles inside the membrane. Make the electrode inside pressure in equilibrium with the atmosphere by opening the electrode refilling aperture during measurement. FAQs: 1. What’s the meaning of electrode condition icon? After calibration, the pH meter will judge the electrode slope and offset, then display the judgment on the screen: slope > 95% Electrode condition is very good slope : 90%-95% Electrode condition is acceptable slope : 85%-90% Electrode condition is not good, need recalibration or replacing. ST270 pH electrode EN-3 2. How do you properly store the pH electrode after every use? A: Rinse the electrode with pure water, put the electrode into the watering cap. If the solution in the cap is polluted or is dried, clean it and fill it with 3M KCl solution. Avoid placing the electrode a dry place or deionized water for a long time. 3. What’s the life of the pH electrodes? A: Measured samples, storage conditions and maintenance methods all could influence the life of pH electrodes; normally a pH electrode expires after 12 months. Some sample, such as strong acid or alkali, corrosive liquid will shorten the pH electrode life. The pH electrode is consumable and is generally provided with warranty. 4. The key operation points for ST270? A: During use, ST270 electrode tip generally does not break. Because of the distance between tip and junction is longer than normal, the depth must exceed this distance, to ensure valid test data. Due to the particularity of samples, ST270 electrode must be cleaned from time to time. For example, when measuring protein medium, we should use the pepsin/HCl cleaning fluid to rinse the electrode. When measuring Oily/organic solution, electrode should be rinsed by acetone or ethanol. EN-4 ST270 pH electrode Troubleshooting If the electrode readings are abnormal, the following procedures should be paid attention to. 1. If the slope is below 95% it needs to be recalibrated. 2. If the measuring depth of electrode head in the sample is low the junction must be in contact with sample. 3. Soak and rinse the electrode in a cleaning solution for 2 hours after use. Electrodo de pH ST270 Modelo ST270 Nºde ítem 30240974 ES-1 Descripción Electrodo de pH de vidrio con punta de lanza 2 en 1 Especificaciones: Conexión: BNC, Cinch Longitud del cable: 1,0 m Longitud del astil: 120 mm Diámetro del astil: 12 mm Rango de pH: 0 a 14 pH Potencial cero: 7,00 ± 0,25 pH Pendiente: ≥ 97% Impedancia: ≤ 100 MΩ Material del astil: Vidrio Electrolito de referencia: Solución 3M KCl Rango de temperatura: 0 a 100 ℃ Sonda de temperatura: NTC 30 kΩ Diafragma de referencia: Manga OHAUS ofrece diferentes electrodos de pH para diferentes aplicaciones. Asegúrese de que ha elegido la opción de electrodo de pH que mejor se adapta a sus muestras, y comience realizando una calibración del electrodo. El electrodo ST270 es un electrodo de pH de vidrio 2 en1. Un electrodo 3 en 1 tiene también una sonda de temperatura. ES-2 Electrodo de pH ST270 Antes de utilizar el electrodo en la calibración o en la medición, tenga en cuenta: 1. Al desembalar el equipo, revise el electrodo para ver si tiene algún daño mecánico en la membrana, en el diafragma o en el cuerpo. 2. Desatornille primero la cubierta de hidratación, luego la tapa, enjuague el electrodo con agua pura. Limpie el electrodo con papel tisú. No frote la membrana de vidrio, ya que el tiempo de respuesta podría aumentar. 3. Elimine las burbujas de aire dentro de la membrana. Equilibre la presión interior del electrodo con la presión atmosférica abriendo la abertura de relleno del electrodo durante la medición. Preguntas frecuentes: 1. ¿Cuál es el significado del icono de condición del electrodo? Después de la calibración, el medidor de pH estima la pendiente y la compensación del electrodo, a continuación, luego muestra la estimación en la pantalla: Pendiente > 95% La condición del electrodo es muy buena Pendiente: 90% a 95% La condición del electrodo es aceptable Pendiente: 85% a 90% La condición del electrodo no es buena, es necesario volver a calibrarlo o sustituirlo. Electrodo de pH ST270 ES-3 2. ¿Cómo se guarda correctamente el electrodo de pH después de cada uso? A: Enjuague el electrodo con agua pura, ponga el electrodo dentro de la cubierta de hidratación. Si la solución en la cubierta está contaminada o está seca, límpiela y llénela con solución 3M KCl. Evite colocar el electrodo en un lugar seco o en agua desionizada durante un largo tiempo. 3. ¿Cuál es la vida útil de los electrodos de pH? A: Las muestras de mediciones, las condiciones de almacenamiento y los métodos de mantenimiento pueden afectar la vida útil de los electrodos de pH; normalmente, la duración de un electrodo de pH expira es de 12 meses. Algunas muestras, tales como ácidos o álcalis fuertes o líquidos corrosivos acortan la vida útil del electrodo de pH. Un electrodo de pH es un artículo desechable y en general se proporciona con garantía. 4. ¿Cuáles son los puntos clave de operación del ST270? A: Durante el uso, la punta del electrodo ST270 en general no se rompe. Debido a que la distancia entre la punta y la juntura es mayor de lo normal, la profundidad debe superar esta distancia, para garantizar la validez de los datos. Debido a la particularidad de las muestras, el electrodo ST270 debe limpiarse de vez en cuando. Por ejemplo, cuando se mide un medio proteico, debe utilizarse un líquido de limpieza a base de pepsina/HCl para limpiar el electrodo. Cuando se mide una solución aceitosa/orgánica, el electrodo debe lavarse con acetona o etanol. ES-4 Electrodo de pH ST270 Solución de problemas Si las lecturas del electrodo son anormales, deben seguirse los siguientes procedimientos. 1. Si la pendiente es menor a 95%, es necesario volver a calibrar. 2. Si la profundidad de medición de la cabeza del electrodo es baja, la juntura debe estar en contacto con la muestra. 3. Remoje y enjuague el electrodo en una solución de limpieza durante 2 horas después del uso. ST270 pH electrode Modèle ST270 Numéro d'article 30240974 FR-1 Description Verre d'électrode pH à pointe effilée 2-en-1 Spécifications: Connexion : BNC, Cinch Longueur de câble : 1.0m La longueur de tiges : 120mm Diamètre de tiges: 12mm Plage de pH : 0-14pH Potentiel zéro : 7,00±0,25 pH Pente : ≥ 97% Impédance : ≤ 100MΩ Le matériau des tiges: Le verre Electrolyte de référence: Solution de KCl 3 M Plage de température: 0-100℃ Sonde de température: NTC 30kΩ Diaphragme de référence: Manchon OHAUS offre électrodes différentes pH pour les applications différentes. Assurez-vous que vous avez choisi le meilleur ajustement électrode pH pour vos échantillons; et commencez par réaliser un étalonnage de l'électrode. ST270 est une électrode pH 2-en-1 en verre. Une électrode de 3-en-1 dispose également d'une sonde de température. FR-2 ST270 électrode pH Avant d'utiliser l'électrode pendant l'étalonnage ou la mesure, veuillez noter : 1. Lors du déballage, vérifiez l'électrode pour détecter tout dommage mécanique de la membrane, le diaphragme ou de l'organisme. 2. Dévisser le capuchon d'arrosage en premier, puis le couvercle, rincez l'électrode avec de l'eau pure. Essuyez l'électrode avec un mouchoir. Ne frottez pas la membrane en verre, ou le temps de réponse pourrait augmenter. 3. Eliminez les bulles d'air à l'intérieur de la membrane. Faîtes l'électrode à l'intérieur à l'équilibre de pression avec l'atmosphère en ouvrant l'électrode remplissant l'aperture au cours de la mesure. Questions fréquemment posées : 1. Quelle est la signification de l'icône de l'état de l'électrode? Après l'étalonnage, l'oxymètre pH jugera l'inclinaison de l'électrode et le décalage, puis affichera l'arrêt sur l'écran : pente > 95 % La condition de sonde est très bonne pente : 90%-95% L'électrode est dans une condition acceptable. pente : 85%-90% La condition de l'électrode n'est pas bonne, il y a besoin d’un étalonnage nouveau ou du remplacement. ST270 pH electrode FR-3 2. Comment pouvez-vous stocker correctement l'électrode pH après chaque utilisation ? A: Rincez l'électrode avec de l'eau pure, placez l'électrode dans le capuchon de mouillage. Si la solution dans le capuchon est polluée ou séchée, nettoyez-le et remplissezle avec 3M solution de KCl. Évitez de placer l'électrode en un endroit sec ou dans l'eau déionisée pour un long moment. 3. Quelle est la durée de vie des électrodes pH? A: Des échantillons mesurés, des méthodes d'entretien et des conditions d'entreposage tous pourraient influencer la vie des électrodes pH; normalement, une électrode pH expire après 12 mois. Certains exemples, tels que l'acide fort ou l'alcalin liquide corrosif raccourciront la durée de vie d'une électrode pH. L'électrode pH est consommable et est généralement fournie avec la garantie. 4. Les points clés de l'opération pour ST270 ? A: Au cours de l'utilisation, type de l'électrode ST270 généralement ne se rompe pas. Étant que la distance entre la pointe et la jonction est plus longue que la normale, la profondeur doit dépasser cette distance, afin de garantir la validité des données du test. En raison de la particularité des échantillons, l'électrode ST270 doit être nettoyée de temps en temps. Par exemple, perdant la mesure de la protéine moyenne, nous devrions utiliser le fluide de nettoyage pepsine/HCl pour rincer l'électrode. Lors de la mesure de la solution organique/huileuse l'électrode doit être rincée avec l'acétone ou l'éthanol. FR-4 ST270 électrode pH Dépannage Si les relevés de l'électrode sont anormaux, vous devez prêter l'attention aux procédures suivantes. 1. Si la pente est inférieure de 95 %, elle a besoin d'être ré-étalonnée. 2. Si la mesure de la profondeur de la tête d'électrode dans l'échantillon est faible la jonction doit être en contact avec l'échantillon. 3. Tremper et rincer l'électrode dans une solution de nettoyage pendant 2 heures après l'utilisation.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14

Ohaus ST270 Manual de usuario

Tipo
Manual de usuario

en otros idiomas