Normative Reference (Appendix)
Immunity Test standard
IEC 60601-1-2 Test Level Compliance Level
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8kV contact
± 2kV, ± 4kV, ± 8kV, ± 15kV air
± 8 kV contact
± 2kV, ± 4kV, ± 8kV, ± 15kV air
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV, for power supply lines
± 1kV signal input/ output
100 kHz repetition frequency
± 2kV, for power supply lines
Not applicable
100 kHz repetition frequency
Surge IEC 61000-4-5
± 0.5kV, ± 1kV differential lines
± 0.5kV, ± 1kV, ± 1kV common mode
± 0.5kV, ± 1kV differential lines
Not Applicable
Voltage dips, interruptions
and variations on power
supply input lines
IEC 61000-4-11
0 % U
T
; 0.5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0 % U
T
; 1 cycle and 70 % U
T
; 25/30
cycles; Single phase: at 0°.
0 % U
T
; 250/300 cycle
0 % U
T
; 0.5 cycle. At 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°.
0 % U
T
; 1 cycle and 70 % U
T
; 25/30
cycles; Single phase: at 0°.
0 % U
T
; 250/300 cycle
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conducted RF
IEC 61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Note: U
is the a.c. mians voltage prior to application of the test level.
WARNING
- Use of this appliance adjacent to or stacked with other device should be avoided because it
could result in improper operation. If such use is necessary, this appliance and the other device
should be observed to verify that they are operating normally. If the appliance does not light on
or ickering, it means that the appliance is not operating normally.
- Don’t near active HF surgical equipment and the RF shielded room of an ME system for
magnetic resonance imaging, where the intensity of EM disturbances is high.
- Do not change any cable without authorization of the manufacturer. Use of cables other
than those specied or provided by the manufacturer of this appliance could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this appliance and result
in improper operation.
- Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm to any part of this appliance, including
power cable. Otherwise, degradation of the performance of this equipment could result.
- If the appliance does not light on or ickering due to the electromagnetic disturbances, do
not use it and try to place the appliance in other location.
- For maintaining the basic safety and essential performance with regard to electromagnetic
disturbances for the expected service life, please precaution to take if the use location is near
(e.g. less than 1,5 km from) from AM, FM or TV broadcast antennas.