150 & 450 Series Waterproof Handheld Meter
pH/mV Operation Instructions
150 & 450 Series Medidor de manoresistente al agua
pH/mV Instrucciones de funcionamiento
Part of Thermo Fisher Scientific Parte de Thermo Fisher Scientific
68X651901 08/14 Rev068X651901 08/14 Rev0
mV Oset Adjustment
1. While in mV measurement mode, dip the ORP and ATC
sensors into a solution with a known mV value (i.e. Zobel,
Light’s, quinhydrone, or iodide/triiodide) and stir.
2. When the “READY indicator appears, use up/down arrows
to adjust the primary reading to match the mV value at
the measured temperature, then press ENTER. The meter
allows an adjustable maximum value of ±200 mV from the
factory default mV value. When an offset has been stored
successfully, R.mV replaces mV.
Temperature Calibration/Manual ATC
1. Press CAL from any measurement, then press MODE.
2. Skip to step 3 for manual ATC, otherwise, dip the temperature
sensor into a solution with a known accurate temperature.
The upper display shows the active temperature while
the lower display shows the factory default temperature
without adjustment.
3. Use up/down arrows to adjust the upper display. Press
ENTER to accept the calibration temperature. The maximum
adjustable value is ±10 ºC (or ±18 ºF) from factory default.
Error Messages
“ERR” will appear when an error condition exists or the
incorrect key is pressed. Common examples include:
• Pressing ENTER during calibration before the READY
indicator appears. Wait for the READY indicator before
pressing ENTER.
UR (Under Range) • OR (Over Range)
Intended Use, Maintenance & Precautions
These handheld meters use sensors to detect various parameters
for water-based measurements. For routine maintenance
disconnect the power cord or battery, then dust or wipe the
display using a damp cloth. If necessary, warm water or a mild
water based detergent can be used. Immediately remove any
spilled substance from contact with the meter using the proper
cleaning procedure for the type of spill.
Do not use this equipment in potentially explosive atmospheres.
Refer to the electrode instructions for use, storage and cleaning.
Ensure that no liquid enters the instrument.
Do not use any aggressive cleaning chemicals (solvents or
similar agents).
• There are no user serviceable parts inside. Attempts to
service internal parts may void the warranty.
WARNING: No modification of this equipment is allowed.
pH Buer Options
Select the desired pH Buffer Calibration Group:
USA (1.68, 4.01, 7.00, 10.01, 12.45) or
NIST (1.68, 4.01, 6.86, 9.18, 12.45) or
DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) or
MAN (manual adjustment of any custom pH values that are
≥1 pH unit apart. 450 series only)
Select number of calibration points
Select Calibration Due Reminder
Set number of days from 0-60 for desired parameter
View Calibration Data
• Press ENTER to view each point that is calibrated.
View Electrode Data
• Press ENTER to view mV Oset and Slope % of the
measured reading.
System Settings
Data Logging:
MANUAL upon key press only
TIMED interval. Choose (SEC / MIN / HOUR) interval.
Automatic shut off after 10 minutes. Choose ON or OFF.
Clock Settings:
Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY).
Time: Choose (24HR or 12HR). If 12HR, choose AM or PM.
Set Printer Type:
CSV (Comma Separated Values) – best format for computer
Printer (Text) – best format for printer.
Choose Manual (MAN) upon key press or TIMED interval.
If timed, choose (SEC / MIN / HOUR).
Reset
NO. Exits from reset menu options without action.
FACTORY RESET. Returns all settings except date/time
and ATC calibration to factory default values after ENTER is
pressed then restarts meter.
DATA RESET. Erases data stored in memory while retaining
other settings after ENTER is pressed.
CALIBRATION RESET. Erases non-ATC calibration data while
retaining other settings after ENTER is pressed.
pH Calibration
For best results, periodic calibration with known, accurate
standards is recommended. Calibrate with standards that
bracket your intended measuring range while including a
neutral standard (pH 7.00 or 6.86). For example, if you expect
to measure samples from pH 6.2 to pH 9.5, calibration with
4.01, 7.00, and 10.01 standards will work well. Provide stirring
for best results. After calibration with two or more points, the
active slope segment of the measurement will be visible on
the bottom display during measurement. 100 % slope will be
shown if only one calibration point is performed and “- - -“ if
no calibration is performed. The meter will automatically return
to measurement mode upon successful completion of the
number of specified calibration points. To specify a different
number of pH calibration points, see pH Buer Options.
Using Automatic Buer Recognition
1. While in pH measurement mode, dip the pH and ATC
sensor(s) into your first standard, then press CAL. The primary
display will search for the nearest standard value, while the
secondary display will show the un-adjusted value.
2. When the “READY indicator appears, press ENTER to
accept. The primary reading will flash DONE”.
3. Rinse your electrode(s) then dip into the next pH standard.
The primary display will search for the nearest standard
value that has not yet been calibrated, while the secondary
display will show the unadjusted value. When the READY
indicator appears, press ENTER to accept.
4. To calibrate another pH standard repeat Step 3 or press
MEAS to return to pH measurement mode.
Using Manual Recognition / Custom Buers
(450 Series Only)
1. While in pH measurement mode, dip the pH and ATC
sensor(s) into your first standard then press CAL.
2. When the READY indicator appears, use up/down arrows
to adjust the primary reading to match the standard value
at the measured temperature, then press ENTER.
3. Rinse your electrode(s) then repeat Step 2 with a standard
that is ≥1 pH unit from the previous standard value.
4. To calibrate another pH standard, repeat Step 3 or press
MEAS to return to pH measurement mode.
pH 150
pH/mV
pH 450
pH/mV/ISE
PC 450
pH/mV/Con
PD 450
pH/mV/DO
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-16/24, Singapore 139949
Tel: (65) 6778-6876
Fax: (65) 6778-0086
eutech@thermofisher.com
www.eutechinst.com
Press once to power ON in the mode that was
previously used. Press again to turn backlight on
for one minute or off (450 series only). Hold for 3
seconds to power OFF.
Toggle between measurement and calibration
modes. In SETUP mode, BACK serves to return to
the previous menu option or setting.
Confirm calibration values in CAL mode.
Confirm selections in SETUP mode.
Freeze or release the measured reading.
Customize instrument settings and preferences.
(See also Setup Programs)
Toggle between available measurement types.
Save measurement into memory. Increase value
or scroll up in SETUP or manual calibration.
Recall saved values from memory. Decrease value
or scroll down in SETUP or manual calibration.
Send output data to printer or computer.
(450 series only).
Models:
Getting Started/Connections
After installing (2) AA batteries and/or connecting the optional
110/220 VAC power supply, connect the desired sensors to the
corresponding ports.
12 mm and 16 mm probes can utilize the Grip-Clip™ to attach
one or more sensors to a beaker and to the instrument as
needed. The stand can be extended as shown above or used
for wall-mounting.
Keypad Functions
Setup Programs
To access the settings below, press SETUP. Up/down arrows
will display the available options. Press ENTER to accept the
desired setting, or BACK to return to the previous option
and/or exit.
Conguration Options
Ready indicator ON / OFF / or Automatic HOLD when stable
• Choose ºCelsius or ºFahrenheit
Oakton Instruments
625 East Bunker Court,
Vernon Hills, IL, 60061, USA
Tel: 1-888-462-5866
Fax: 1-847-247-2984
www.4oakton.com
DO: 9-Pin RDO/ATC
(PD 450 only)
BNC: pH,
mV/ORP or
Ion Selective
ATC/Con:
6-Pin ATC or
ATC/Conductivity
(PC 450 only)
A/C Power
Adapter
Output (USB or
RS232 – PC or Printer
Communication)
Instrument Operating Conditions
Operating Ambient Temp. 5 to 45 ºC
Operating Relative Humidity 5 to 85 %, non-condensing
Storage Temp. -20 to +60 ºC
Storage Relative Humidity 5 to 85 %, non-condensing
Pollution Degree 2
Overvoltage Category II
Weight 500 g
Size (L x W x H) 21.15 x 9.87 x 5.85 cm
Regulatory & Safety CE, TUV 3-1, FCC Class A
Power Rating DC Input: 9 VDC 1 A
Battery Requirement
2 x AA (LR6) 1.5 V batteries
(replace batteries when
battery sign blinks)
Vibration Shipping/handling per ISTA #1A
Shock
Drop test in packaging per
ISTA #1A
Enclosure (Designed To Meet) IP67 (using rubber covers)
Universal Power Adapter Operating Conditions
Operating Ambient Temp. 0 to 50 ºC
Operating Relative Humidity 0 to 90 %, non-condensing
Storage Temp. -20 to +75 ºC
Storage Relative Humidity 0 to 90 %, non-condensing
Pollution Degree 2
Overvoltage Category II
Power Rating
I/P: 100 - 240 V, 50/60 Hz, 0.3A
O/P: 9 VDC 1 A
2. Cuando aparezca el indicador READY (Listo) use las flechas arriba/
abajo para ajustar la lectura primaria para ajustar el valor mV a la
temperatura medida, luego oprima ENTER. El medidor permite un valor
máximo ajustable de ±200 mV a partir del valor mV estándar de fábrica.
Cuando un offset se haya guardado exitosamente, mV es reeplazado
por R.mV.
Calibración de temperatura/ATC Manuel
1. Oprima CAL de cualquier medición, luego oprima MODE (Modo).
2. Siga con el paso 3 por una ATC manual, de otra manera sumerja el sensor
de temperatura en una solución con una temperatura determinada
conocida. La visualización superior muestra la temperatura activa
mientras que la visualización inferior muestra la temperatura estándar
de fábrica sin ajuste.
3. Use las flechas arriba/abajo para ajustar la visualización superior. Oprima
ENTER para aceptar la temperatura de calibración. El valor ajustable
máximo es ±10 ºC (o ±18 ºF) desde el valor de fábrica.
Mensajes de error
“ERR” va a aparecer si existe una condición de error o si se presionada
la tecla equivocada. Ejemplos comunes son entre otros:
• Apretar ENTER durante la calibración antes de que aparezca la indicación
READY (Listo). Espere a que aparezca la indicación READY (Listo) antes
de apretar ENTER.
RI (Rango inferior) • RS (Rango superior)
Uso intencionado, mantenimiento y precauciones
Estos medidores portables utilizan sensores para detectar varios
parámetros para mediciones basadas en agua. Para el mantenimiento de
rutina disconecte el cable de alimentación o la batería y luego pase un
trapo húmedo por la pantalla. Si es necesario puede utilizarse agua tibia o
un detergente suave basado en agua. Remueve inmediatamente cualquier
substancia derramada sobre el medidor con el procedimiento de limpieza
correcto para el tipo de substancia derramada.
No utilice este equipo en atmósferas potencialmente peligrosas.
Vea las instrucciones del electrodo para su uso, almacenamiento y
limpieza.
Asegúrese de que ningún liquido ingrese al instrumento.
• No use químicos de limpieza agresivos (solventes o substancias
similares).
No hay partes en el interior que pueden ser mantenidos por el usuario.
El intento de mantener partes internas puede anular la garantía.
ADVERTENCIA: No está permitido ninguna modificación de este equipo.
NIST (1.68, 4.01, 6.86, 9.18, 12.45) or
DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) o
MAN (ajuste manual de valores de ph personalizados que están a ≥1
unidad de pH. Sóloserie 450)
Seleccione la cantidad de puntos de calibración
Seleccione Noticación de Calibración Pendiente
Fije la cantidad de días entre 0 y 60 para el parámetro deseado
Ver Datos de Calibración
• Presione ENTER para ver cada punto que va a ser calibrado.
Ver Datos de Electrodos
• Presione ENTER para ver el oset de mV y el % de pendiente (Slope) del
valor medido.
Ajustes del Sistema
Registro de datos:
MANUAL al apretar un botónsólo intervalo TIMED (Cronometrado). Elija
(SEG / MIN / HORA) intervalo.
Apagado automático después de 10 minutos. Elija ON oder OFF.
Ajustes del reloj:
Fecha: Elija USA (MM/DD/AAAA) o Euro (DD/MM/AAAA).
Tiempo: Elija (24HR o 12HR). Si 12HR elija AM o PM.
Ajustartipo de impresora:
CSV (Valores Separados por Coma) – mejor formato para la computadora.
Impresora (texto) – mejor formato para impresora.
Elija Manual (MAN) al apretar un botón o intervalo TIMED
(Cronometrado).
Si cronometrado, elija (SEG / MIN / HORA).
Resetear
NO (NRO). de salidas de las opciones del menú de reseteo sin acción.
FACTORY RESET (RESETEO DE FÁBRICA). Retorna todos los ajustes von
excepción de fecha/tiempo y calibración ATC a los valores estándares de
fábrica luego de oprimir ENTER y el medidor es reseteado.
DATA RESET (RESETEO DE DATOS). Elimina datos guardados en la
memoria mientras otros ajustes se conservan luego de oprimir ENTER.
CALIBRATION RESET (RESETEO DE CALIBRACIÓN). Elimina datos de
calibración no-ATC mientras otros ajustes se conservan luego de oprimir
ENTER.
Calibración de pH
Para mejores resultados se recomienda una calibración periódica con
estándares conocidos y precisos. Calibre con estándares que encuadren su
rango de medición deseado mientras incluye un estándar neutro (pH 7,00
o 6,86). Por ejemplo si usted espera medir muestras con un pH de entre 6,2
y 9,5 la calibración con los estándares 4,01, 7,00 y 10,01 van a funcionar bien.
Revuelva para los mejores resultados. Luego de la calibración con dos o
más puntos el segmento de pendiente activo de la medición va a ser visible
en la visualización inferior durante la medición. Una pendiente de 100 % va
a aparecer si solamente se utiliza un punto de calibración y “- - -“ si no se
lleva a cabo ninguna calibración. El medidor va volver automáticamente
al modo de medición luego de una finalización exitosa de la cantidad de
puntos de calibración especificada. Para especificar una cantidad distinta
de puntos de calibración de pH vea Opciones de Buer de pH.
Utilizando Reconocimiento Automático de Buer
1. Mientras esté en el modo de medición pH sumerja los sensores de
pH y ATC en su primer estándar, luego presione CAL. La visualización
primaria va a buscar al valor estándar más cercano mientras que la
pantalla secundaria va a mostrar el valor no ajustado.
2. Cuando aparece el indicador READY (Listo) presione ENTER para
aceptar. En la medición primaria va a brillar READY (Listo).
3. Lave sus electrodos y sumérjalos en el siguiente estándar de pH. La
visualización primaria va a buscar al valor estándar más cercano, que
no ha sido calibrado aún, mientras que la visualización secundaria va
a mostrar el valor no ajustado. Cuando aparece el indicador READY
(Listo) presione ENTER para aceptar.
4. Para calibrar otro estándar de pH, repita el paso 3 o presione MEAS, para
regresar al modo de medición de pH.
Utilizar Reconocimiento Manual / Buers Personalizados (Sólo
Serie 450)
1. Mientras esté en el modo de medición pH sumerja los sensores de pH y
ATC en su primer estándar, luego presione CAL.
2. Cuando aparezca el indicador READY (Listo) utilize las flechas arriba/
abajo para ajustar la medición primaria para que sea igual que el valor
estándar a la temperatura medida, luego presione ENTER.
3. Lave sus electrodos y luego repita el paso 2 con un estándar que sea ≥1
unidad de pH a partir del valor estándar anterior.
4. Para calibrar otro estándar de pH, repita el paso 3 o presione MEAS, para
regresar al modo de medición de pH.
Ajuste del oset de mV
1. En el modo de medición mV sumerja los sensores ORP y ATZ en una
solución con un valor mV conocido (z. B. Zobel, Light, quinhydrona o
ioduro/triioduro) y revuelva.
pH 150
pH/mV
pH 450
pH/mV/ISE
PC 450
pH/mV/Con
PD 450
pH/mV/DO
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-16/24, Singapore 139949
Tel: (65) 6778-6876
Fax: (65) 6778-0086
eutech@thermofisher.com
www.eutechinst.com
Presione una vez, para encender en el modo usado
previamente. Presione otra vez, para encender la
retroiluminación por un minuto o apagar (sólo serie 450).
Mantenga apretado por 3 segundos.
Alterne ente los modos medición y calibración. En el modo
CONFIGURACIÓN la tecla VOLVER sirve para volver a la
opción previa del menú o ajuste.
Confirmar los valores de calibración en el modo CAL.
Confirmar selecciones en el modo CONFIGURACIÓN.
Congelar o liberar el valor medido.
Personalizar los ajustes del instrumento y las preferencias.
(Vea también los programas de conguración)
Alterne entre los tipos de medición disponibles.
Guardar la medición en la memoria. Aumente el valor o
desplace hacia arriba en CONFIGURACIÓN o calibración
manual.
Volver a mostrar valores guardados en la memoria.
Reduzca el valor o desplace hacia abajo en
CONFIGURACIÓN o calibración manual.
Envíe los datos de salida a una impresora o computadora.
(Sólo serie 450)
Modelos:
Primeros pasos/Conexiones
Luego de insertar (2) baterías AA y/o conectar el suministro de energía
opcional de 110/220 VAC conecte los sensores deseados a los puertos
correspondientes.
Sondas de 12 mm y 16 mm pueden utilizar Grip-Clip™ para sujetar uno o
más sensores a un vaso y si es necesario al instrumento. El pie puede ser
extendido como se muestra arriba o puede ser montado a la pared.
Funciones del teclado
Programmes de conguration
Para acceder a los ajustes abajo presione CONFIGURACIÓN, Las flechas
arriba/abajo van a hacer aparecer las opciones disponibles. Presione ENTER
para aceptar el ajuste deseado o BACK (Volver) para volver a la opción
anterior y/o salir.
Opciones de Conguración
Indicador listo ON / OFF / o HOLD (Mantenimiento) automático si
estable
• Elija ºCelsius o ºFahrenheit
Opciones de buer de pH
Seleccione el grupo de calibración del buffer de pH deseado:
USA (1.68, 4.01, 7.00, 10.01, 12.45) or
Oakton Instruments
625 East Bunker Court,
Vernon Hills, IL, 60061, USA
Tel: 1-888-462-5866
Fax: 1-847-247-2984
www.4oakton.com
DO: 9-Pin RDO/ATC
(sólo PD 450)
BNC: pH, mV/ORP
o ATC/Cond
selectiva de iones
ATC/Con:
6-Pin ATC o
ATC/Conductividad
(sólo PC 450) DO
Transformador
A/C
Salida (USB o RS232 –
Comunicación de
PC o Impresora)
Condiciones de funcionamiento del instrumento
Temperatura de ambiente de
funcionamiento
5 a 45 ºC
Humedad relativa de
funcionamiento
5 a 85 %, no condensada
Temperatura de almacenamiento -20 a +60 ºC
Humedad relativa de
almacenamiento
5 a 85 %, no condensada
Contaminación Grado de 2
Sobretensión Categoría de II
Peso 500 g
Tamaño (L x A x A) 21,15 x 9,87 x 5,85 cm
Ordenanzas y Seguridad CE, TUV 3-1, FCC Class A
Potencia Nominal Entrada de CC: 9 VDC 1 A
Requerimientos de batería
2 x AA (LR6) 1,5 V batterías
(reemplazar batterías cuando
parpadea el cartel de la batería)
Vibración Envío/Manejo conforme a ISTA #1A
Shock
Test de caída en envase conforme
a ISTA #1A
Carcasa (diseñado para cumplir) IP67 (utilizando cubiertas de goma)
Transformador universal Condiciones de funcionamiento
Temperatura de Ambiente de
funcionamiento
5 a 50 ºC
Humedad relativa de
funcionamiento
0 a 90 %, no condensada
Temperatura de almacenamiento -20 a +75 ºC
Humedad relativa de
almacenamiento
0 a 90 %, no condensada
Contaminación Grado de 2
Sobretensión Categoría de II
Potencia Nominal
I/P: 100 - 240 V, 50/60 Hz, 0.3A
O/P: 9 VDC 1 A
150 & 450 Series Waterproof Handheld Meter
pH/mV Operation Instructions
150 & 450 Series Medidor de manoresistente al agua
pH/mV Instrucciones de funcionamiento
Part of Thermo Fisher Scientific Parte de Thermo Fisher Scientific
68X651901 08/14 Rev068X651901 08/14 Rev0
mV Oset Adjustment
1. While in mV measurement mode, dip the ORP and ATC
sensors into a solution with a known mV value (i.e. Zobel,
Light’s, quinhydrone, or iodide/triiodide) and stir.
2. When the “READY indicator appears, use up/down arrows
to adjust the primary reading to match the mV value at
the measured temperature, then press ENTER. The meter
allows an adjustable maximum value of ±200 mV from the
factory default mV value. When an offset has been stored
successfully, R.mV replaces mV.
Temperature Calibration/Manual ATC
1. Press CAL from any measurement, then press MODE.
2. Skip to step 3 for manual ATC, otherwise, dip the temperature
sensor into a solution with a known accurate temperature.
The upper display shows the active temperature while
the lower display shows the factory default temperature
without adjustment.
3. Use up/down arrows to adjust the upper display. Press
ENTER to accept the calibration temperature. The maximum
adjustable value is ±10 ºC (or ±18 ºF) from factory default.
Error Messages
“ERR” will appear when an error condition exists or the
incorrect key is pressed. Common examples include:
• Pressing ENTER during calibration before the READY
indicator appears. Wait for the READY indicator before
pressing ENTER.
UR (Under Range) • OR (Over Range)
Intended Use, Maintenance & Precautions
These handheld meters use sensors to detect various parameters
for water-based measurements. For routine maintenance
disconnect the power cord or battery, then dust or wipe the
display using a damp cloth. If necessary, warm water or a mild
water based detergent can be used. Immediately remove any
spilled substance from contact with the meter using the proper
cleaning procedure for the type of spill.
Do not use this equipment in potentially explosive atmospheres.
Refer to the electrode instructions for use, storage and cleaning.
Ensure that no liquid enters the instrument.
Do not use any aggressive cleaning chemicals (solvents or
similar agents).
• There are no user serviceable parts inside. Attempts to
service internal parts may void the warranty.
WARNING: No modification of this equipment is allowed.
pH Buer Options
Select the desired pH Buffer Calibration Group:
USA (1.68, 4.01, 7.00, 10.01, 12.45) or
NIST (1.68, 4.01, 6.86, 9.18, 12.45) or
DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) or
MAN (manual adjustment of any custom pH values that are
≥1 pH unit apart. 450 series only)
Select number of calibration points
Select Calibration Due Reminder
Set number of days from 0-60 for desired parameter
View Calibration Data
• Press ENTER to view each point that is calibrated.
View Electrode Data
• Press ENTER to view mV Oset and Slope % of the
measured reading.
System Settings
Data Logging:
MANUAL upon key press only
TIMED interval. Choose (SEC / MIN / HOUR) interval.
Automatic shut off after 10 minutes. Choose ON or OFF.
Clock Settings:
Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY).
Time: Choose (24HR or 12HR). If 12HR, choose AM or PM.
Set Printer Type:
CSV (Comma Separated Values) – best format for computer
Printer (Text) – best format for printer.
Choose Manual (MAN) upon key press or TIMED interval.
If timed, choose (SEC / MIN / HOUR).
Reset
NO. Exits from reset menu options without action.
FACTORY RESET. Returns all settings except date/time
and ATC calibration to factory default values after ENTER is
pressed then restarts meter.
DATA RESET. Erases data stored in memory while retaining
other settings after ENTER is pressed.
CALIBRATION RESET. Erases non-ATC calibration data while
retaining other settings after ENTER is pressed.
pH Calibration
For best results, periodic calibration with known, accurate
standards is recommended. Calibrate with standards that
bracket your intended measuring range while including a
neutral standard (pH 7.00 or 6.86). For example, if you expect
to measure samples from pH 6.2 to pH 9.5, calibration with
4.01, 7.00, and 10.01 standards will work well. Provide stirring
for best results. After calibration with two or more points, the
active slope segment of the measurement will be visible on
the bottom display during measurement. 100 % slope will be
shown if only one calibration point is performed and “- - -“ if
no calibration is performed. The meter will automatically return
to measurement mode upon successful completion of the
number of specified calibration points. To specify a different
number of pH calibration points, see pH Buer Options.
Using Automatic Buer Recognition
1. While in pH measurement mode, dip the pH and ATC
sensor(s) into your first standard, then press CAL. The primary
display will search for the nearest standard value, while the
secondary display will show the un-adjusted value.
2. When the “READY indicator appears, press ENTER to
accept. The primary reading will flash DONE”.
3. Rinse your electrode(s) then dip into the next pH standard.
The primary display will search for the nearest standard
value that has not yet been calibrated, while the secondary
display will show the unadjusted value. When the READY
indicator appears, press ENTER to accept.
4. To calibrate another pH standard repeat Step 3 or press
MEAS to return to pH measurement mode.
Using Manual Recognition / Custom Buers
(450 Series Only)
1. While in pH measurement mode, dip the pH and ATC
sensor(s) into your first standard then press CAL.
2. When the READY indicator appears, use up/down arrows
to adjust the primary reading to match the standard value
at the measured temperature, then press ENTER.
3. Rinse your electrode(s) then repeat Step 2 with a standard
that is ≥1 pH unit from the previous standard value.
4. To calibrate another pH standard, repeat Step 3 or press
MEAS to return to pH measurement mode.
pH 150
pH/mV
pH 450
pH/mV/ISE
PC 450
pH/mV/Con
PD 450
pH/mV/DO
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-16/24, Singapore 139949
Tel: (65) 6778-6876
Fax: (65) 6778-0086
eutech@thermofisher.com
www.eutechinst.com
Press once to power ON in the mode that was
previously used. Press again to turn backlight on
for one minute or off (450 series only). Hold for 3
seconds to power OFF.
Toggle between measurement and calibration
modes. In SETUP mode, BACK serves to return to
the previous menu option or setting.
Confirm calibration values in CAL mode.
Confirm selections in SETUP mode.
Freeze or release the measured reading.
Customize instrument settings and preferences.
(See also Setup Programs)
Toggle between available measurement types.
Save measurement into memory. Increase value
or scroll up in SETUP or manual calibration.
Recall saved values from memory. Decrease value
or scroll down in SETUP or manual calibration.
Send output data to printer or computer.
(450 series only).
Models:
Getting Started/Connections
After installing (2) AA batteries and/or connecting the optional
110/220 VAC power supply, connect the desired sensors to the
corresponding ports.
12 mm and 16 mm probes can utilize the Grip-Clip™ to attach
one or more sensors to a beaker and to the instrument as
needed. The stand can be extended as shown above or used
for wall-mounting.
Keypad Functions
Setup Programs
To access the settings below, press SETUP. Up/down arrows
will display the available options. Press ENTER to accept the
desired setting, or BACK to return to the previous option
and/or exit.
Conguration Options
Ready indicator ON / OFF / or Automatic HOLD when stable
• Choose ºCelsius or ºFahrenheit
Oakton Instruments
625 East Bunker Court,
Vernon Hills, IL, 60061, USA
Tel: 1-888-462-5866
Fax: 1-847-247-2984
www.4oakton.com
DO: 9-Pin RDO/ATC
(PD 450 only)
BNC: pH,
mV/ORP or
Ion Selective
ATC/Con:
6-Pin ATC or
ATC/Conductivity
(PC 450 only)
A/C Power
Adapter
Output (USB or
RS232 – PC or Printer
Communication)
Instrument Operating Conditions
Operating Ambient Temp. 5 to 45 ºC
Operating Relative Humidity 5 to 85 %, non-condensing
Storage Temp. -20 to +60 ºC
Storage Relative Humidity 5 to 85 %, non-condensing
Pollution Degree 2
Overvoltage Category II
Weight 500 g
Size (L x W x H) 21.15 x 9.87 x 5.85 cm
Regulatory & Safety CE, TUV 3-1, FCC Class A
Power Rating DC Input: 9 VDC 1 A
Battery Requirement
2 x AA (LR6) 1.5 V batteries
(replace batteries when
battery sign blinks)
Vibration Shipping/handling per ISTA #1A
Shock
Drop test in packaging per
ISTA #1A
Enclosure (Designed To Meet) IP67 (using rubber covers)
Universal Power Adapter Operating Conditions
Operating Ambient Temp. 0 to 50 ºC
Operating Relative Humidity 0 to 90 %, non-condensing
Storage Temp. -20 to +75 ºC
Storage Relative Humidity 0 to 90 %, non-condensing
Pollution Degree 2
Overvoltage Category II
Power Rating
I/P: 100 - 240 V, 50/60 Hz, 0.3A
O/P: 9 VDC 1 A
2. Cuando aparezca el indicador READY (Listo) use las flechas arriba/
abajo para ajustar la lectura primaria para ajustar el valor mV a la
temperatura medida, luego oprima ENTER. El medidor permite un valor
máximo ajustable de ±200 mV a partir del valor mV estándar de fábrica.
Cuando un offset se haya guardado exitosamente, mV es reeplazado
por R.mV.
Calibración de temperatura/ATC Manuel
1. Oprima CAL de cualquier medición, luego oprima MODE (Modo).
2. Siga con el paso 3 por una ATC manual, de otra manera sumerja el sensor
de temperatura en una solución con una temperatura determinada
conocida. La visualización superior muestra la temperatura activa
mientras que la visualización inferior muestra la temperatura estándar
de fábrica sin ajuste.
3. Use las flechas arriba/abajo para ajustar la visualización superior. Oprima
ENTER para aceptar la temperatura de calibración. El valor ajustable
máximo es ±10 ºC (o ±18 ºF) desde el valor de fábrica.
Mensajes de error
“ERR” va a aparecer si existe una condición de error o si se presionada
la tecla equivocada. Ejemplos comunes son entre otros:
• Apretar ENTER durante la calibración antes de que aparezca la indicación
READY (Listo). Espere a que aparezca la indicación READY (Listo) antes
de apretar ENTER.
RI (Rango inferior) • RS (Rango superior)
Uso intencionado, mantenimiento y precauciones
Estos medidores portables utilizan sensores para detectar varios
parámetros para mediciones basadas en agua. Para el mantenimiento de
rutina disconecte el cable de alimentación o la batería y luego pase un
trapo húmedo por la pantalla. Si es necesario puede utilizarse agua tibia o
un detergente suave basado en agua. Remueve inmediatamente cualquier
substancia derramada sobre el medidor con el procedimiento de limpieza
correcto para el tipo de substancia derramada.
No utilice este equipo en atmósferas potencialmente peligrosas.
Vea las instrucciones del electrodo para su uso, almacenamiento y
limpieza.
Asegúrese de que ningún liquido ingrese al instrumento.
• No use químicos de limpieza agresivos (solventes o substancias
similares).
No hay partes en el interior que pueden ser mantenidos por el usuario.
El intento de mantener partes internas puede anular la garantía.
ADVERTENCIA: No está permitido ninguna modificación de este equipo.
NIST (1.68, 4.01, 6.86, 9.18, 12.45) or
DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) o
MAN (ajuste manual de valores de ph personalizados que están a ≥1
unidad de pH. Sóloserie 450)
Seleccione la cantidad de puntos de calibración
Seleccione Noticación de Calibración Pendiente
Fije la cantidad de días entre 0 y 60 para el parámetro deseado
Ver Datos de Calibración
• Presione ENTER para ver cada punto que va a ser calibrado.
Ver Datos de Electrodos
• Presione ENTER para ver el oset de mV y el % de pendiente (Slope) del
valor medido.
Ajustes del Sistema
Registro de datos:
MANUAL al apretar un botónsólo intervalo TIMED (Cronometrado). Elija
(SEG / MIN / HORA) intervalo.
Apagado automático después de 10 minutos. Elija ON oder OFF.
Ajustes del reloj:
Fecha: Elija USA (MM/DD/AAAA) o Euro (DD/MM/AAAA).
Tiempo: Elija (24HR o 12HR). Si 12HR elija AM o PM.
Ajustartipo de impresora:
CSV (Valores Separados por Coma) – mejor formato para la computadora.
Impresora (texto) – mejor formato para impresora.
Elija Manual (MAN) al apretar un botón o intervalo TIMED
(Cronometrado).
Si cronometrado, elija (SEG / MIN / HORA).
Resetear
NO (NRO). de salidas de las opciones del menú de reseteo sin acción.
FACTORY RESET (RESETEO DE FÁBRICA). Retorna todos los ajustes von
excepción de fecha/tiempo y calibración ATC a los valores estándares de
fábrica luego de oprimir ENTER y el medidor es reseteado.
DATA RESET (RESETEO DE DATOS). Elimina datos guardados en la
memoria mientras otros ajustes se conservan luego de oprimir ENTER.
CALIBRATION RESET (RESETEO DE CALIBRACIÓN). Elimina datos de
calibración no-ATC mientras otros ajustes se conservan luego de oprimir
ENTER.
Calibración de pH
Para mejores resultados se recomienda una calibración periódica con
estándares conocidos y precisos. Calibre con estándares que encuadren su
rango de medición deseado mientras incluye un estándar neutro (pH 7,00
o 6,86). Por ejemplo si usted espera medir muestras con un pH de entre 6,2
y 9,5 la calibración con los estándares 4,01, 7,00 y 10,01 van a funcionar bien.
Revuelva para los mejores resultados. Luego de la calibración con dos o
más puntos el segmento de pendiente activo de la medición va a ser visible
en la visualización inferior durante la medición. Una pendiente de 100 % va
a aparecer si solamente se utiliza un punto de calibración y “- - -“ si no se
lleva a cabo ninguna calibración. El medidor va volver automáticamente
al modo de medición luego de una finalización exitosa de la cantidad de
puntos de calibración especificada. Para especificar una cantidad distinta
de puntos de calibración de pH vea Opciones de Buer de pH.
Utilizando Reconocimiento Automático de Buer
1. Mientras esté en el modo de medición pH sumerja los sensores de
pH y ATC en su primer estándar, luego presione CAL. La visualización
primaria va a buscar al valor estándar más cercano mientras que la
pantalla secundaria va a mostrar el valor no ajustado.
2. Cuando aparece el indicador READY (Listo) presione ENTER para
aceptar. En la medición primaria va a brillar READY (Listo).
3. Lave sus electrodos y sumérjalos en el siguiente estándar de pH. La
visualización primaria va a buscar al valor estándar más cercano, que
no ha sido calibrado aún, mientras que la visualización secundaria va
a mostrar el valor no ajustado. Cuando aparece el indicador READY
(Listo) presione ENTER para aceptar.
4. Para calibrar otro estándar de pH, repita el paso 3 o presione MEAS, para
regresar al modo de medición de pH.
Utilizar Reconocimiento Manual / Buers Personalizados (Sólo
Serie 450)
1. Mientras esté en el modo de medición pH sumerja los sensores de pH y
ATC en su primer estándar, luego presione CAL.
2. Cuando aparezca el indicador READY (Listo) utilize las flechas arriba/
abajo para ajustar la medición primaria para que sea igual que el valor
estándar a la temperatura medida, luego presione ENTER.
3. Lave sus electrodos y luego repita el paso 2 con un estándar que sea ≥1
unidad de pH a partir del valor estándar anterior.
4. Para calibrar otro estándar de pH, repita el paso 3 o presione MEAS, para
regresar al modo de medición de pH.
Ajuste del oset de mV
1. En el modo de medición mV sumerja los sensores ORP y ATZ en una
solución con un valor mV conocido (z. B. Zobel, Light, quinhydrona o
ioduro/triioduro) y revuelva.
pH 150
pH/mV
pH 450
pH/mV/ISE
PC 450
pH/mV/Con
PD 450
pH/mV/DO
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-16/24, Singapore 139949
Tel: (65) 6778-6876
Fax: (65) 6778-0086
eutech@thermofisher.com
www.eutechinst.com
Presione una vez, para encender en el modo usado
previamente. Presione otra vez, para encender la
retroiluminación por un minuto o apagar (sólo serie 450).
Mantenga apretado por 3 segundos.
Alterne ente los modos medición y calibración. En el modo
CONFIGURACIÓN la tecla VOLVER sirve para volver a la
opción previa del menú o ajuste.
Confirmar los valores de calibración en el modo CAL.
Confirmar selecciones en el modo CONFIGURACIÓN.
Congelar o liberar el valor medido.
Personalizar los ajustes del instrumento y las preferencias.
(Vea también los programas de conguración)
Alterne entre los tipos de medición disponibles.
Guardar la medición en la memoria. Aumente el valor o
desplace hacia arriba en CONFIGURACIÓN o calibración
manual.
Volver a mostrar valores guardados en la memoria.
Reduzca el valor o desplace hacia abajo en
CONFIGURACIÓN o calibración manual.
Envíe los datos de salida a una impresora o computadora.
(Sólo serie 450)
Modelos:
Primeros pasos/Conexiones
Luego de insertar (2) baterías AA y/o conectar el suministro de energía
opcional de 110/220 VAC conecte los sensores deseados a los puertos
correspondientes.
Sondas de 12 mm y 16 mm pueden utilizar Grip-Clip™ para sujetar uno o
más sensores a un vaso y si es necesario al instrumento. El pie puede ser
extendido como se muestra arriba o puede ser montado a la pared.
Funciones del teclado
Programmes de conguration
Para acceder a los ajustes abajo presione CONFIGURACIÓN, Las flechas
arriba/abajo van a hacer aparecer las opciones disponibles. Presione ENTER
para aceptar el ajuste deseado o BACK (Volver) para volver a la opción
anterior y/o salir.
Opciones de Conguración
Indicador listo ON / OFF / o HOLD (Mantenimiento) automático si
estable
• Elija ºCelsius o ºFahrenheit
Opciones de buer de pH
Seleccione el grupo de calibración del buffer de pH deseado:
USA (1.68, 4.01, 7.00, 10.01, 12.45) or
Oakton Instruments
625 East Bunker Court,
Vernon Hills, IL, 60061, USA
Tel: 1-888-462-5866
Fax: 1-847-247-2984
www.4oakton.com
DO: 9-Pin RDO/ATC
(sólo PD 450)
BNC: pH, mV/ORP
o ATC/Cond
selectiva de iones
ATC/Con:
6-Pin ATC o
ATC/Conductividad
(sólo PC 450) DO
Transformador
A/C
Salida (USB o RS232 –
Comunicación de
PC o Impresora)
Condiciones de funcionamiento del instrumento
Temperatura de ambiente de
funcionamiento
5 a 45 ºC
Humedad relativa de
funcionamiento
5 a 85 %, no condensada
Temperatura de almacenamiento -20 a +60 ºC
Humedad relativa de
almacenamiento
5 a 85 %, no condensada
Contaminación Grado de 2
Sobretensión Categoría de II
Peso 500 g
Tamaño (L x A x A) 21,15 x 9,87 x 5,85 cm
Ordenanzas y Seguridad CE, TUV 3-1, FCC Class A
Potencia Nominal Entrada de CC: 9 VDC 1 A
Requerimientos de batería
2 x AA (LR6) 1,5 V batterías
(reemplazar batterías cuando
parpadea el cartel de la batería)
Vibración Envío/Manejo conforme a ISTA #1A
Shock
Test de caída en envase conforme
a ISTA #1A
Carcasa (diseñado para cumplir) IP67 (utilizando cubiertas de goma)
Transformador universal Condiciones de funcionamiento
Temperatura de Ambiente de
funcionamiento
5 a 50 ºC
Humedad relativa de
funcionamiento
0 a 90 %, no condensada
Temperatura de almacenamiento -20 a +75 ºC
Humedad relativa de
almacenamiento
0 a 90 %, no condensada
Contaminación Grado de 2
Sobretensión Categoría de II
Potencia Nominal
I/P: 100 - 240 V, 50/60 Hz, 0.3A
O/P: 9 VDC 1 A

Transcripción de documentos

Part of Thermo Fisher Scientific 150 & 450 Series Waterproof Handheld Meter 150 & 450 Series Medidor de manoresistente al agua pH/mV Operation Instructions Models: pH 150 pH/mV pH 450 pH/mV/ISE PC 450 pH/mV/Con PD 450 pH/mV/DO Getting Started/Connections After installing (2) AA batteries and/or connecting the optional 110/220 VAC power supply, connect the desired sensors to the corresponding ports. DO: 9-Pin RDO/ATC (PD 450 only) BNC: pH, mV/ORP or Ion Selective ATC/Con: 6-Pin ATC or ATC/Conductivity (PC 450 only) A/C Power Adapter Output (USB or RS232 – PC or Printer Communication) 12 mm and 16 mm probes can utilize the Grip-Clip™ to attach one or more sensors to a beaker and to the instrument as needed. The stand can be extended as shown above or used for wall-mounting. Keypad Functions Press once to power ON in the mode that was previously used. Press again to turn backlight on for one minute or off (450 series only). Hold for 3 seconds to power OFF. Toggle between measurement and calibration modes. In SETUP mode, BACK serves to return to the previous menu option or setting. Confirm calibration values in CAL mode. Confirm selections in SETUP mode. Freeze or release the measured reading. Customize instrument settings and preferences. (See also Setup Programs) Toggle between available measurement types. Save measurement into memory. Increase value or scroll up in SETUP or manual calibration. Recall saved values from memory. Decrease value or scroll down in SETUP or manual calibration. Send output data to printer or computer. (450 series only). Setup Programs To access the settings below, press SETUP. Up/down arrows will display the available options. Press ENTER to accept the desired setting, or BACK to return to the previous option and/or exit. Configuration Options • Ready indicator ON / OFF / or Automatic HOLD when stable • Choose ºCelsius or ºFahrenheit pH Buffer Options • Select the desired pH Buffer Calibration Group: USA (1.68, 4.01, 7.00, 10.01, 12.45) or NIST (1.68, 4.01, 6.86, 9.18, 12.45) or DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) or MAN (manual adjustment of any custom pH values that are ≥1 pH unit apart. 450 series only) • Select number of calibration points Select Calibration Due Reminder • Set number of days from 0-60 for desired parameter View Calibration Data • Press ENTER to view each point that is calibrated. View Electrode Data • Press ENTER to view mV Offset and Slope % of the measured reading. System Settings • Data Logging: MANUAL upon key press only TIMED interval. Choose (SEC / MIN / HOUR) interval. • Automatic shut off after 10 minutes. Choose ON or OFF. • Clock Settings: Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY). Time: Choose (24HR or 12HR). If 12HR, choose AM or PM. • Set Printer Type: CSV (Comma Separated Values) – best format for computer Printer (Text) – best format for printer. Choose Manual (MAN) upon key press or TIMED interval. If timed, choose (SEC / MIN / HOUR). Reset • NO. Exits from reset menu options without action. • FACTORY RESET. Returns all settings except date/time and ATC calibration to factory default values after ENTER is pressed then restarts meter. • DATA RESET. Erases data stored in memory while retaining other settings after ENTER is pressed. • CALIBRATION RESET. Erases non-ATC calibration data while retaining other settings after ENTER is pressed. pH Calibration For best results, periodic calibration with known, accurate standards is recommended. Calibrate with standards that bracket your intended measuring range while including a neutral standard (pH 7.00 or 6.86). For example, if you expect to measure samples from pH 6.2 to pH 9.5, calibration with 4.01, 7.00, and 10.01 standards will work well. Provide stirring for best results. After calibration with two or more points, the active slope segment of the measurement will be visible on the bottom display during measurement. 100 % slope will be shown if only one calibration point is performed and “- - -“ if no calibration is performed. The meter will automatically return to measurement mode upon successful completion of the number of specified calibration points. To specify a different number of pH calibration points, see pH Buffer Options. Using Automatic Buffer Recognition 1. While in pH measurement mode, dip the pH and ATC sensor(s) into your first standard, then press CAL. The primary display will search for the nearest standard value, while the secondary display will show the un-adjusted value. 2. When the “READY” indicator appears, press ENTER to accept. The primary reading will flash “DONE”. 3. Rinse your electrode(s) then dip into the next pH standard. The primary display will search for the nearest standard value that has not yet been calibrated, while the secondary display will show the unadjusted value. When the “READY” indicator appears, press ENTER to accept. 4. To calibrate another pH standard repeat Step 3 or press MEAS to return to pH measurement mode. Using Manual Recognition / Custom Buffers (450 Series Only) 1. While in pH measurement mode, dip the pH and ATC sensor(s) into your first standard then press CAL. 2. When the READY indicator appears, use up/down arrows to adjust the primary reading to match the standard value at the measured temperature, then press ENTER. 3. Rinse your electrode(s) then repeat Step 2 with a standard that is ≥1 pH unit from the previous standard value. 4. To calibrate another pH standard, repeat Step 3 or press MEAS to return to pH measurement mode. Parte de Thermo Fisher Scientific pH/mV Instrucciones de funcionamiento Modelos: mV Offset Adjustment 1. While in mV measurement mode, dip the ORP and ATC sensors into a solution with a known mV value (i.e. Zobel, Light’s, quinhydrone, or iodide/triiodide) and stir. 2. When the “READY” indicator appears, use up/down arrows to adjust the primary reading to match the mV value at the measured temperature, then press ENTER. The meter allows an adjustable maximum value of ±200 mV from the factory default mV value. When an offset has been stored successfully, R.mV replaces mV. Temperature Calibration/Manual ATC 1. Press CAL from any measurement, then press MODE. 2. Skip to step 3 for manual ATC, otherwise, dip the temperature sensor into a solution with a known accurate temperature. The upper display shows the active temperature while the lower display shows the factory default temperature without adjustment. 3. Use up/down arrows to adjust the upper display. Press ENTER to accept the calibration temperature. The maximum adjustable value is ±10 ºC (or ±18 ºF) from factory default. Error Messages pH 150 pH/mV pH 450 pH/mV/ISE Primeros pasos/Conexiones DO: 9-Pin RDO/ATC (sólo PD 450) Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24, Singapore 139949 Tel: (65) 6778-6876 Fax: (65) 6778-0086 [email protected] www.eutechinst.com Oakton Instruments 625 East Bunker Court, Vernon Hills, IL, 60061, USA Tel: 1-888-462-5866 Fax: 1-847-247-2984 [email protected] www.4oakton.com 68X651901 08/14 Rev0 BNC: pH, mV/ORP o ATC/Cond selectiva de iones ATC/Con: 6-Pin ATC o ATC/Conductividad (sólo PC 450) DO Intended Use, Maintenance & Precautions Instrument Operating Conditions Operating Ambient Temp. 5 to 45 ºC Operating Relative Humidity 5 to 85 %, non-condensing Storage Temp. -20 to +60 ºC Storage Relative Humidity 5 to 85 %, non-condensing Pollution Degree 2 Overvoltage Category II Weight 500 g Size (L x W x H) 21.15 x 9.87 x 5.85 cm Regulatory & Safety CE, TUV 3-1, FCC Class A Power Rating DC Input: 9 VDC 1 A 2 x AA (LR6) 1.5 V batteries Battery Requirement (replace batteries when battery sign blinks) Vibration Shipping/handling per ISTA #1A Drop test in packaging per Shock ISTA #1A Enclosure (Designed To Meet) IP67 (using rubber covers) Universal Power Adapter Operating Conditions Operating Ambient Temp. 0 to 50 ºC Operating Relative Humidity 0 to 90 %, non-condensing Storage Temp. -20 to +75 ºC Storage Relative Humidity 0 to 90 %, non-condensing Pollution Degree 2 Overvoltage Category II I/P: 100 - 240 V, 50/60 Hz, 0.3A Power Rating O/P: 9 VDC 1 A PD 450 pH/mV/DO Luego de insertar (2) baterías AA y/o conectar el suministro de energía opcional de 110/220 VAC conecte los sensores deseados a los puertos correspondientes. “ERR” will appear when an error condition exists or the incorrect key is pressed. Common examples include: • Pressing ENTER during calibration before the “READY” indicator appears. Wait for the “READY” indicator before pressing ENTER. • UR (Under Range) • OR (Over Range) These handheld meters use sensors to detect various parameters for water-based measurements. For routine maintenance disconnect the power cord or battery, then dust or wipe the display using a damp cloth. If necessary, warm water or a mild water based detergent can be used. Immediately remove any spilled substance from contact with the meter using the proper cleaning procedure for the type of spill. • Do not use this equipment in potentially explosive atmospheres. • Refer to the electrode instructions for use, storage and cleaning. • Ensure that no liquid enters the instrument. • Do not use any aggressive cleaning chemicals (solvents or similar agents). • There are no user serviceable parts inside. Attempts to service internal parts may void the warranty. • WARNING: No modification of this equipment is allowed. PC 450 pH/mV/Con Transformador A/C Salida (USB o RS232 – Comunicación de PC o Impresora) Sondas de 12 mm y 16 mm pueden utilizar Grip-Clip™ para sujetar uno o más sensores a un vaso y si es necesario al instrumento. El pie puede ser extendido como se muestra arriba o puede ser montado a la pared. Funciones del teclado Presione una vez, para encender en el modo usado previamente. Presione otra vez, para encender la retroiluminación por un minuto o apagar (sólo serie 450). Mantenga apretado por 3 segundos. Alterne ente los modos medición y calibración. En el modo CONFIGURACIÓN la tecla VOLVER sirve para volver a la opción previa del menú o ajuste. Confirmar los valores de calibración en el modo CAL. Confirmar selecciones en el modo CONFIGURACIÓN. Congelar o liberar el valor medido. Personalizar los ajustes del instrumento y las preferencias. (Vea también los programas de configuración) Alterne entre los tipos de medición disponibles. Guardar la medición en la memoria. Aumente el valor o desplace hacia arriba en CONFIGURACIÓN o calibración manual. Volver a mostrar valores guardados en la memoria. Reduzca el valor o desplace hacia abajo en CONFIGURACIÓN o calibración manual. Envíe los datos de salida a una impresora o computadora. (Sólo serie 450) Programmes de configuration Para acceder a los ajustes abajo presione CONFIGURACIÓN, Las flechas arriba/abajo van a hacer aparecer las opciones disponibles. Presione ENTER para aceptar el ajuste deseado o BACK (Volver) para volver a la opción anterior y/o salir. Opciones de Configuración • Indicador listo ON / OFF / o HOLD (Mantenimiento) automático si estable • Elija ºCelsius o ºFahrenheit Opciones de buffer de pH • Seleccione el grupo de calibración del buffer de pH deseado: USA (1.68, 4.01, 7.00, 10.01, 12.45) or NIST (1.68, 4.01, 6.86, 9.18, 12.45) or DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) o MAN (ajuste manual de valores de ph personalizados que están a ≥1 unidad de pH. Sóloserie 450) • Seleccione la cantidad de puntos de calibración Seleccione Notificación de Calibración Pendiente • Fije la cantidad de días entre 0 y 60 para el parámetro deseado Ver Datos de Calibración • Presione ENTER para ver cada punto que va a ser calibrado. Ver Datos de Electrodos • Presione ENTER para ver el offset de mV y el % de pendiente (Slope) del valor medido. Ajustes del Sistema • Registro de datos: MANUAL al apretar un botónsólo intervalo TIMED (Cronometrado). Elija (SEG / MIN / HORA) intervalo. • Apagado automático después de 10 minutos. Elija ON oder OFF. • Ajustes del reloj: Fecha: Elija USA (MM/DD/AAAA) o Euro (DD/MM/AAAA). Tiempo: Elija (24HR o 12HR). Si 12HR elija AM o PM. • Ajustartipo de impresora: CSV (Valores Separados por Coma) – mejor formato para la computadora. Impresora (texto) – mejor formato para impresora. Elija Manual (MAN) al apretar un botón o intervalo TIMED (Cronometrado). Si cronometrado, elija (SEG / MIN / HORA). Resetear • NO (NRO). de salidas de las opciones del menú de reseteo sin acción. • FACTORY RESET (RESETEO DE FÁBRICA). Retorna todos los ajustes von excepción de fecha/tiempo y calibración ATC a los valores estándares de fábrica luego de oprimir ENTER y el medidor es reseteado. • DATA RESET (RESETEO DE DATOS). Elimina datos guardados en la memoria mientras otros ajustes se conservan luego de oprimir ENTER. • CALIBRATION RESET (RESETEO DE CALIBRACIÓN). Elimina datos de calibración no-ATC mientras otros ajustes se conservan luego de oprimir ENTER. Calibración de pH Para mejores resultados se recomienda una calibración periódica con estándares conocidos y precisos. Calibre con estándares que encuadren su rango de medición deseado mientras incluye un estándar neutro (pH 7,00 o 6,86). Por ejemplo si usted espera medir muestras con un pH de entre 6,2 y 9,5 la calibración con los estándares 4,01, 7,00 y 10,01 van a funcionar bien. Revuelva para los mejores resultados. Luego de la calibración con dos o más puntos el segmento de pendiente activo de la medición va a ser visible en la visualización inferior durante la medición. Una pendiente de 100 % va a aparecer si solamente se utiliza un punto de calibración y “- - -“ si no se lleva a cabo ninguna calibración. El medidor va volver automáticamente al modo de medición luego de una finalización exitosa de la cantidad de puntos de calibración especificada. Para especificar una cantidad distinta de puntos de calibración de pH vea Opciones de Buffer de pH. Utilizando Reconocimiento Automático de Buffer 1. Mientras esté en el modo de medición pH sumerja los sensores de pH y ATC en su primer estándar, luego presione CAL. La visualización primaria va a buscar al valor estándar más cercano mientras que la pantalla secundaria va a mostrar el valor no ajustado. 2. Cuando aparece el indicador “READY” (Listo) presione ENTER para aceptar. En la medición primaria va a brillar “READY” (Listo). 3. Lave sus electrodos y sumérjalos en el siguiente estándar de pH. La visualización primaria va a buscar al valor estándar más cercano, que no ha sido calibrado aún, mientras que la visualización secundaria va a mostrar el valor no ajustado. Cuando aparece el indicador “READY” (Listo) presione ENTER para aceptar. 4. Para calibrar otro estándar de pH, repita el paso 3 o presione MEAS, para regresar al modo de medición de pH. Utilizar Reconocimiento Manual / Buffers Personalizados (Sólo Serie 450) 1. Mientras esté en el modo de medición pH sumerja los sensores de pH y ATC en su primer estándar, luego presione CAL. 2. Cuando aparezca el indicador “READY” (Listo) utilize las flechas arriba/ abajo para ajustar la medición primaria para que sea igual que el valor estándar a la temperatura medida, luego presione ENTER. 3. Lave sus electrodos y luego repita el paso 2 con un estándar que sea ≥1 unidad de pH a partir del valor estándar anterior. 4. Para calibrar otro estándar de pH, repita el paso 3 o presione MEAS, para regresar al modo de medición de pH. Ajuste del offset de mV 1. En el modo de medición mV sumerja los sensores ORP y ATZ en una solución con un valor mV conocido (z. B. Zobel, Light, quinhydrona o ioduro/triioduro) y revuelva. 2. Cuando aparezca el indicador “READY” (Listo) use las flechas arriba/ abajo para ajustar la lectura primaria para ajustar el valor mV a la temperatura medida, luego oprima ENTER. El medidor permite un valor máximo ajustable de ±200 mV a partir del valor mV estándar de fábrica. Cuando un offset se haya guardado exitosamente, mV es reeplazado por R.mV. Calibración de temperatura/ATC Manuel 1. Oprima CAL de cualquier medición, luego oprima MODE (Modo). 2. Siga con el paso 3 por una ATC manual, de otra manera sumerja el sensor de temperatura en una solución con una temperatura determinada conocida. La visualización superior muestra la temperatura activa mientras que la visualización inferior muestra la temperatura estándar de fábrica sin ajuste. 3. Use las flechas arriba/abajo para ajustar la visualización superior. Oprima ENTER para aceptar la temperatura de calibración. El valor ajustable máximo es ±10 ºC (o ±18 ºF) desde el valor de fábrica. Mensajes de error “ERR” va a aparecer si existe una condición de error o si se presionada la tecla equivocada. Ejemplos comunes son entre otros: • Apretar ENTER durante la calibración antes de que aparezca la indicación “READY” (Listo). Espere a que aparezca la indicación “READY” (Listo) antes de apretar ENTER. • RI (Rango inferior) • RS (Rango superior) Uso intencionado, mantenimiento y precauciones Estos medidores portables utilizan sensores para detectar varios parámetros para mediciones basadas en agua. Para el mantenimiento de rutina disconecte el cable de alimentación o la batería y luego pase un trapo húmedo por la pantalla. Si es necesario puede utilizarse agua tibia o un detergente suave basado en agua. Remueve inmediatamente cualquier substancia derramada sobre el medidor con el procedimiento de limpieza correcto para el tipo de substancia derramada. • No utilice este equipo en atmósferas potencialmente peligrosas. • Vea las instrucciones del electrodo para su uso, almacenamiento y limpieza. • Asegúrese de que ningún liquido ingrese al instrumento. • No use químicos de limpieza agresivos (solventes o substancias similares). • No hay partes en el interior que pueden ser mantenidos por el usuario. El intento de mantener partes internas puede anular la garantía. • ADVERTENCIA: No está permitido ninguna modificación de este equipo. Condiciones de funcionamiento del instrumento Temperatura de ambiente de 5 a 45 ºC funcionamiento Humedad relativa de 5 a 85 %, no condensada funcionamiento Temperatura de almacenamiento -20 a +60 ºC Humedad relativa de 5 a 85 %, no condensada almacenamiento Contaminación Sobretensión Peso Tamaño (L x A x A) Ordenanzas y Seguridad Potencia Nominal Grado de 2 Categoría de II 500 g 21,15 x 9,87 x 5,85 cm CE, TUV 3-1, FCC Class A Entrada de CC: 9 VDC 1 A 2 x AA (LR6) 1,5 V batterías Requerimientos de batería (reemplazar batterías cuando parpadea el cartel de la batería) Vibración Envío/Manejo conforme a ISTA #1A Test de caída en envase conforme Shock a ISTA #1A Carcasa (diseñado para cumplir) IP67 (utilizando cubiertas de goma) Transformador universal Condiciones de funcionamiento Temperatura de Ambiente de 5 a 50 ºC funcionamiento Humedad relativa de 0 a 90 %, no condensada funcionamiento Temperatura de almacenamiento -20 a +75 ºC Humedad relativa de 0 a 90 %, no condensada almacenamiento Contaminación Grado de 2 Sobretensión Categoría de II I/P: 100 - 240 V, 50/60 Hz, 0.3A Potencia Nominal O/P: 9 VDC 1 A Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24, Singapore 139949 Tel: (65) 6778-6876 Fax: (65) 6778-0086 [email protected] www.eutechinst.com Oakton Instruments 625 East Bunker Court, Vernon Hills, IL, 60061, USA Tel: 1-888-462-5866 Fax: 1-847-247-2984 [email protected] www.4oakton.com 68X651901 08/14 Rev0 Part of Thermo Fisher Scientific 150 & 450 Series Waterproof Handheld Meter 150 & 450 Series Medidor de manoresistente al agua pH/mV Operation Instructions Models: pH 150 pH/mV pH 450 pH/mV/ISE PC 450 pH/mV/Con PD 450 pH/mV/DO Getting Started/Connections After installing (2) AA batteries and/or connecting the optional 110/220 VAC power supply, connect the desired sensors to the corresponding ports. DO: 9-Pin RDO/ATC (PD 450 only) BNC: pH, mV/ORP or Ion Selective ATC/Con: 6-Pin ATC or ATC/Conductivity (PC 450 only) A/C Power Adapter Output (USB or RS232 – PC or Printer Communication) 12 mm and 16 mm probes can utilize the Grip-Clip™ to attach one or more sensors to a beaker and to the instrument as needed. The stand can be extended as shown above or used for wall-mounting. Keypad Functions Press once to power ON in the mode that was previously used. Press again to turn backlight on for one minute or off (450 series only). Hold for 3 seconds to power OFF. Toggle between measurement and calibration modes. In SETUP mode, BACK serves to return to the previous menu option or setting. Confirm calibration values in CAL mode. Confirm selections in SETUP mode. Freeze or release the measured reading. Customize instrument settings and preferences. (See also Setup Programs) Toggle between available measurement types. Save measurement into memory. Increase value or scroll up in SETUP or manual calibration. Recall saved values from memory. Decrease value or scroll down in SETUP or manual calibration. Send output data to printer or computer. (450 series only). Setup Programs To access the settings below, press SETUP. Up/down arrows will display the available options. Press ENTER to accept the desired setting, or BACK to return to the previous option and/or exit. Configuration Options • Ready indicator ON / OFF / or Automatic HOLD when stable • Choose ºCelsius or ºFahrenheit pH Buffer Options • Select the desired pH Buffer Calibration Group: USA (1.68, 4.01, 7.00, 10.01, 12.45) or NIST (1.68, 4.01, 6.86, 9.18, 12.45) or DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) or MAN (manual adjustment of any custom pH values that are ≥1 pH unit apart. 450 series only) • Select number of calibration points Select Calibration Due Reminder • Set number of days from 0-60 for desired parameter View Calibration Data • Press ENTER to view each point that is calibrated. View Electrode Data • Press ENTER to view mV Offset and Slope % of the measured reading. System Settings • Data Logging: MANUAL upon key press only TIMED interval. Choose (SEC / MIN / HOUR) interval. • Automatic shut off after 10 minutes. Choose ON or OFF. • Clock Settings: Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY). Time: Choose (24HR or 12HR). If 12HR, choose AM or PM. • Set Printer Type: CSV (Comma Separated Values) – best format for computer Printer (Text) – best format for printer. Choose Manual (MAN) upon key press or TIMED interval. If timed, choose (SEC / MIN / HOUR). Reset • NO. Exits from reset menu options without action. • FACTORY RESET. Returns all settings except date/time and ATC calibration to factory default values after ENTER is pressed then restarts meter. • DATA RESET. Erases data stored in memory while retaining other settings after ENTER is pressed. • CALIBRATION RESET. Erases non-ATC calibration data while retaining other settings after ENTER is pressed. pH Calibration For best results, periodic calibration with known, accurate standards is recommended. Calibrate with standards that bracket your intended measuring range while including a neutral standard (pH 7.00 or 6.86). For example, if you expect to measure samples from pH 6.2 to pH 9.5, calibration with 4.01, 7.00, and 10.01 standards will work well. Provide stirring for best results. After calibration with two or more points, the active slope segment of the measurement will be visible on the bottom display during measurement. 100 % slope will be shown if only one calibration point is performed and “- - -“ if no calibration is performed. The meter will automatically return to measurement mode upon successful completion of the number of specified calibration points. To specify a different number of pH calibration points, see pH Buffer Options. Using Automatic Buffer Recognition 1. While in pH measurement mode, dip the pH and ATC sensor(s) into your first standard, then press CAL. The primary display will search for the nearest standard value, while the secondary display will show the un-adjusted value. 2. When the “READY” indicator appears, press ENTER to accept. The primary reading will flash “DONE”. 3. Rinse your electrode(s) then dip into the next pH standard. The primary display will search for the nearest standard value that has not yet been calibrated, while the secondary display will show the unadjusted value. When the “READY” indicator appears, press ENTER to accept. 4. To calibrate another pH standard repeat Step 3 or press MEAS to return to pH measurement mode. Using Manual Recognition / Custom Buffers (450 Series Only) 1. While in pH measurement mode, dip the pH and ATC sensor(s) into your first standard then press CAL. 2. When the READY indicator appears, use up/down arrows to adjust the primary reading to match the standard value at the measured temperature, then press ENTER. 3. Rinse your electrode(s) then repeat Step 2 with a standard that is ≥1 pH unit from the previous standard value. 4. To calibrate another pH standard, repeat Step 3 or press MEAS to return to pH measurement mode. Parte de Thermo Fisher Scientific pH/mV Instrucciones de funcionamiento Modelos: mV Offset Adjustment 1. While in mV measurement mode, dip the ORP and ATC sensors into a solution with a known mV value (i.e. Zobel, Light’s, quinhydrone, or iodide/triiodide) and stir. 2. When the “READY” indicator appears, use up/down arrows to adjust the primary reading to match the mV value at the measured temperature, then press ENTER. The meter allows an adjustable maximum value of ±200 mV from the factory default mV value. When an offset has been stored successfully, R.mV replaces mV. Temperature Calibration/Manual ATC 1. Press CAL from any measurement, then press MODE. 2. Skip to step 3 for manual ATC, otherwise, dip the temperature sensor into a solution with a known accurate temperature. The upper display shows the active temperature while the lower display shows the factory default temperature without adjustment. 3. Use up/down arrows to adjust the upper display. Press ENTER to accept the calibration temperature. The maximum adjustable value is ±10 ºC (or ±18 ºF) from factory default. Error Messages pH 150 pH/mV pH 450 pH/mV/ISE Primeros pasos/Conexiones DO: 9-Pin RDO/ATC (sólo PD 450) Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24, Singapore 139949 Tel: (65) 6778-6876 Fax: (65) 6778-0086 [email protected] www.eutechinst.com Oakton Instruments 625 East Bunker Court, Vernon Hills, IL, 60061, USA Tel: 1-888-462-5866 Fax: 1-847-247-2984 [email protected] www.4oakton.com 68X651901 08/14 Rev0 BNC: pH, mV/ORP o ATC/Cond selectiva de iones ATC/Con: 6-Pin ATC o ATC/Conductividad (sólo PC 450) DO Intended Use, Maintenance & Precautions Instrument Operating Conditions Operating Ambient Temp. 5 to 45 ºC Operating Relative Humidity 5 to 85 %, non-condensing Storage Temp. -20 to +60 ºC Storage Relative Humidity 5 to 85 %, non-condensing Pollution Degree 2 Overvoltage Category II Weight 500 g Size (L x W x H) 21.15 x 9.87 x 5.85 cm Regulatory & Safety CE, TUV 3-1, FCC Class A Power Rating DC Input: 9 VDC 1 A 2 x AA (LR6) 1.5 V batteries Battery Requirement (replace batteries when battery sign blinks) Vibration Shipping/handling per ISTA #1A Drop test in packaging per Shock ISTA #1A Enclosure (Designed To Meet) IP67 (using rubber covers) Universal Power Adapter Operating Conditions Operating Ambient Temp. 0 to 50 ºC Operating Relative Humidity 0 to 90 %, non-condensing Storage Temp. -20 to +75 ºC Storage Relative Humidity 0 to 90 %, non-condensing Pollution Degree 2 Overvoltage Category II I/P: 100 - 240 V, 50/60 Hz, 0.3A Power Rating O/P: 9 VDC 1 A PD 450 pH/mV/DO Luego de insertar (2) baterías AA y/o conectar el suministro de energía opcional de 110/220 VAC conecte los sensores deseados a los puertos correspondientes. “ERR” will appear when an error condition exists or the incorrect key is pressed. Common examples include: • Pressing ENTER during calibration before the “READY” indicator appears. Wait for the “READY” indicator before pressing ENTER. • UR (Under Range) • OR (Over Range) These handheld meters use sensors to detect various parameters for water-based measurements. For routine maintenance disconnect the power cord or battery, then dust or wipe the display using a damp cloth. If necessary, warm water or a mild water based detergent can be used. Immediately remove any spilled substance from contact with the meter using the proper cleaning procedure for the type of spill. • Do not use this equipment in potentially explosive atmospheres. • Refer to the electrode instructions for use, storage and cleaning. • Ensure that no liquid enters the instrument. • Do not use any aggressive cleaning chemicals (solvents or similar agents). • There are no user serviceable parts inside. Attempts to service internal parts may void the warranty. • WARNING: No modification of this equipment is allowed. PC 450 pH/mV/Con Transformador A/C Salida (USB o RS232 – Comunicación de PC o Impresora) Sondas de 12 mm y 16 mm pueden utilizar Grip-Clip™ para sujetar uno o más sensores a un vaso y si es necesario al instrumento. El pie puede ser extendido como se muestra arriba o puede ser montado a la pared. Funciones del teclado Presione una vez, para encender en el modo usado previamente. Presione otra vez, para encender la retroiluminación por un minuto o apagar (sólo serie 450). Mantenga apretado por 3 segundos. Alterne ente los modos medición y calibración. En el modo CONFIGURACIÓN la tecla VOLVER sirve para volver a la opción previa del menú o ajuste. Confirmar los valores de calibración en el modo CAL. Confirmar selecciones en el modo CONFIGURACIÓN. Congelar o liberar el valor medido. Personalizar los ajustes del instrumento y las preferencias. (Vea también los programas de configuración) Alterne entre los tipos de medición disponibles. Guardar la medición en la memoria. Aumente el valor o desplace hacia arriba en CONFIGURACIÓN o calibración manual. Volver a mostrar valores guardados en la memoria. Reduzca el valor o desplace hacia abajo en CONFIGURACIÓN o calibración manual. Envíe los datos de salida a una impresora o computadora. (Sólo serie 450) Programmes de configuration Para acceder a los ajustes abajo presione CONFIGURACIÓN, Las flechas arriba/abajo van a hacer aparecer las opciones disponibles. Presione ENTER para aceptar el ajuste deseado o BACK (Volver) para volver a la opción anterior y/o salir. Opciones de Configuración • Indicador listo ON / OFF / o HOLD (Mantenimiento) automático si estable • Elija ºCelsius o ºFahrenheit Opciones de buffer de pH • Seleccione el grupo de calibración del buffer de pH deseado: USA (1.68, 4.01, 7.00, 10.01, 12.45) or NIST (1.68, 4.01, 6.86, 9.18, 12.45) or DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) o MAN (ajuste manual de valores de ph personalizados que están a ≥1 unidad de pH. Sóloserie 450) • Seleccione la cantidad de puntos de calibración Seleccione Notificación de Calibración Pendiente • Fije la cantidad de días entre 0 y 60 para el parámetro deseado Ver Datos de Calibración • Presione ENTER para ver cada punto que va a ser calibrado. Ver Datos de Electrodos • Presione ENTER para ver el offset de mV y el % de pendiente (Slope) del valor medido. Ajustes del Sistema • Registro de datos: MANUAL al apretar un botónsólo intervalo TIMED (Cronometrado). Elija (SEG / MIN / HORA) intervalo. • Apagado automático después de 10 minutos. Elija ON oder OFF. • Ajustes del reloj: Fecha: Elija USA (MM/DD/AAAA) o Euro (DD/MM/AAAA). Tiempo: Elija (24HR o 12HR). Si 12HR elija AM o PM. • Ajustartipo de impresora: CSV (Valores Separados por Coma) – mejor formato para la computadora. Impresora (texto) – mejor formato para impresora. Elija Manual (MAN) al apretar un botón o intervalo TIMED (Cronometrado). Si cronometrado, elija (SEG / MIN / HORA). Resetear • NO (NRO). de salidas de las opciones del menú de reseteo sin acción. • FACTORY RESET (RESETEO DE FÁBRICA). Retorna todos los ajustes von excepción de fecha/tiempo y calibración ATC a los valores estándares de fábrica luego de oprimir ENTER y el medidor es reseteado. • DATA RESET (RESETEO DE DATOS). Elimina datos guardados en la memoria mientras otros ajustes se conservan luego de oprimir ENTER. • CALIBRATION RESET (RESETEO DE CALIBRACIÓN). Elimina datos de calibración no-ATC mientras otros ajustes se conservan luego de oprimir ENTER. Calibración de pH Para mejores resultados se recomienda una calibración periódica con estándares conocidos y precisos. Calibre con estándares que encuadren su rango de medición deseado mientras incluye un estándar neutro (pH 7,00 o 6,86). Por ejemplo si usted espera medir muestras con un pH de entre 6,2 y 9,5 la calibración con los estándares 4,01, 7,00 y 10,01 van a funcionar bien. Revuelva para los mejores resultados. Luego de la calibración con dos o más puntos el segmento de pendiente activo de la medición va a ser visible en la visualización inferior durante la medición. Una pendiente de 100 % va a aparecer si solamente se utiliza un punto de calibración y “- - -“ si no se lleva a cabo ninguna calibración. El medidor va volver automáticamente al modo de medición luego de una finalización exitosa de la cantidad de puntos de calibración especificada. Para especificar una cantidad distinta de puntos de calibración de pH vea Opciones de Buffer de pH. Utilizando Reconocimiento Automático de Buffer 1. Mientras esté en el modo de medición pH sumerja los sensores de pH y ATC en su primer estándar, luego presione CAL. La visualización primaria va a buscar al valor estándar más cercano mientras que la pantalla secundaria va a mostrar el valor no ajustado. 2. Cuando aparece el indicador “READY” (Listo) presione ENTER para aceptar. En la medición primaria va a brillar “READY” (Listo). 3. Lave sus electrodos y sumérjalos en el siguiente estándar de pH. La visualización primaria va a buscar al valor estándar más cercano, que no ha sido calibrado aún, mientras que la visualización secundaria va a mostrar el valor no ajustado. Cuando aparece el indicador “READY” (Listo) presione ENTER para aceptar. 4. Para calibrar otro estándar de pH, repita el paso 3 o presione MEAS, para regresar al modo de medición de pH. Utilizar Reconocimiento Manual / Buffers Personalizados (Sólo Serie 450) 1. Mientras esté en el modo de medición pH sumerja los sensores de pH y ATC en su primer estándar, luego presione CAL. 2. Cuando aparezca el indicador “READY” (Listo) utilize las flechas arriba/ abajo para ajustar la medición primaria para que sea igual que el valor estándar a la temperatura medida, luego presione ENTER. 3. Lave sus electrodos y luego repita el paso 2 con un estándar que sea ≥1 unidad de pH a partir del valor estándar anterior. 4. Para calibrar otro estándar de pH, repita el paso 3 o presione MEAS, para regresar al modo de medición de pH. Ajuste del offset de mV 1. En el modo de medición mV sumerja los sensores ORP y ATZ en una solución con un valor mV conocido (z. B. Zobel, Light, quinhydrona o ioduro/triioduro) y revuelva. 2. Cuando aparezca el indicador “READY” (Listo) use las flechas arriba/ abajo para ajustar la lectura primaria para ajustar el valor mV a la temperatura medida, luego oprima ENTER. El medidor permite un valor máximo ajustable de ±200 mV a partir del valor mV estándar de fábrica. Cuando un offset se haya guardado exitosamente, mV es reeplazado por R.mV. Calibración de temperatura/ATC Manuel 1. Oprima CAL de cualquier medición, luego oprima MODE (Modo). 2. Siga con el paso 3 por una ATC manual, de otra manera sumerja el sensor de temperatura en una solución con una temperatura determinada conocida. La visualización superior muestra la temperatura activa mientras que la visualización inferior muestra la temperatura estándar de fábrica sin ajuste. 3. Use las flechas arriba/abajo para ajustar la visualización superior. Oprima ENTER para aceptar la temperatura de calibración. El valor ajustable máximo es ±10 ºC (o ±18 ºF) desde el valor de fábrica. Mensajes de error “ERR” va a aparecer si existe una condición de error o si se presionada la tecla equivocada. Ejemplos comunes son entre otros: • Apretar ENTER durante la calibración antes de que aparezca la indicación “READY” (Listo). Espere a que aparezca la indicación “READY” (Listo) antes de apretar ENTER. • RI (Rango inferior) • RS (Rango superior) Uso intencionado, mantenimiento y precauciones Estos medidores portables utilizan sensores para detectar varios parámetros para mediciones basadas en agua. Para el mantenimiento de rutina disconecte el cable de alimentación o la batería y luego pase un trapo húmedo por la pantalla. Si es necesario puede utilizarse agua tibia o un detergente suave basado en agua. Remueve inmediatamente cualquier substancia derramada sobre el medidor con el procedimiento de limpieza correcto para el tipo de substancia derramada. • No utilice este equipo en atmósferas potencialmente peligrosas. • Vea las instrucciones del electrodo para su uso, almacenamiento y limpieza. • Asegúrese de que ningún liquido ingrese al instrumento. • No use químicos de limpieza agresivos (solventes o substancias similares). • No hay partes en el interior que pueden ser mantenidos por el usuario. El intento de mantener partes internas puede anular la garantía. • ADVERTENCIA: No está permitido ninguna modificación de este equipo. Condiciones de funcionamiento del instrumento Temperatura de ambiente de 5 a 45 ºC funcionamiento Humedad relativa de 5 a 85 %, no condensada funcionamiento Temperatura de almacenamiento -20 a +60 ºC Humedad relativa de 5 a 85 %, no condensada almacenamiento Contaminación Sobretensión Peso Tamaño (L x A x A) Ordenanzas y Seguridad Potencia Nominal Grado de 2 Categoría de II 500 g 21,15 x 9,87 x 5,85 cm CE, TUV 3-1, FCC Class A Entrada de CC: 9 VDC 1 A 2 x AA (LR6) 1,5 V batterías Requerimientos de batería (reemplazar batterías cuando parpadea el cartel de la batería) Vibración Envío/Manejo conforme a ISTA #1A Test de caída en envase conforme Shock a ISTA #1A Carcasa (diseñado para cumplir) IP67 (utilizando cubiertas de goma) Transformador universal Condiciones de funcionamiento Temperatura de Ambiente de 5 a 50 ºC funcionamiento Humedad relativa de 0 a 90 %, no condensada funcionamiento Temperatura de almacenamiento -20 a +75 ºC Humedad relativa de 0 a 90 %, no condensada almacenamiento Contaminación Grado de 2 Sobretensión Categoría de II I/P: 100 - 240 V, 50/60 Hz, 0.3A Potencia Nominal O/P: 9 VDC 1 A Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24, Singapore 139949 Tel: (65) 6778-6876 Fax: (65) 6778-0086 [email protected] www.eutechinst.com Oakton Instruments 625 East Bunker Court, Vernon Hills, IL, 60061, USA Tel: 1-888-462-5866 Fax: 1-847-247-2984 [email protected] www.4oakton.com 68X651901 08/14 Rev0
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

Oakton WD-35632-80 El manual del propietario

Categoría
Medir, probar
Tipo
El manual del propietario