C61wd G4 Lumens Cleaning System (LCS) Installation Guide and Operator’s Manual Copyright 2015 SciCan Ltd. All rights reserved.
LCS Adaptor and Rack
(included)
LCS Plug
(included)
Rack for vertical instruments (optional)
HYDRIM
®
C61wd G4 Lumens Cleaning System (LCS)
Connecting handpieces to the LCS adaptor
Using the touchscreen menu
Screw the LCS fitting
to the appropriate
handpiece coupling.
Insert the handpiece coupling into the handpiece. Push the handpiece into one
of the ports.
The HYDRIM will
automatically detect
the presence of the
LCS adaptor. Select
the desired program
and follow the prompts.
Upon completion, open
the door and remove
the handpieces from the
adaptor by pushing on the
tab at the side of the port.
Example: Regular cycle
1 2 3
The Lumens Cleaning System
(LCS) Adaptor can only be
used with HYDRIM C61wd G4
models equipped with a lumens
cleaning system. If your HYDRIM
is LCS enabled, you will see a
white plug on the inside back
left of the chamber.
1 2 3
English
Multiflex
Intramatic
Push tab
to release
Setting up the LCS unit
Remove the screw and
pull out the white plug on
the back wall of the unit’s
chamber.
Pull the quick connect
fitting ring back, insert air
line coupling and release
the ring to lock.
Attach the LCS plug to the
HYDRIM trolley. (Stays in
the down position when
LCS rack is not being used)
Attach the LCS adaptor and rack
to the trolley, engaging the tab at
the top of the LCS plug. Place the
cannula cleaning basket on the rails
of the LCS rack.
1 2 43
Please note: As recommended
by leading manufacturer/suppliers
of handpieces, the hollow parts
of instruments should be dried
immediately after cleaning and
disinfection using medical compressed
air before additional processes or
sterilization. This is to be followed by
recommended regular maintenance
procedures and application of
maintenance agents/oils. Please
follow all applicable health & safety
regulations for your region.
Maintenance and troubleshooting
LCS system Messages What it Means What to Do – User
CLEANING THE LCS WATER FILTER
REPLACING THE BACTERIOLOGICAL AIR FILTER
Return the filter to the
housing and close the
cap tightly.
Remove the filter. Rinse the filter and, if required,
clean with a soft brush or using
compressed air. Check and clean
the O-ring. Remove any debris
inside the housing.
Unscrew the cap at
the front of the
LCS adaptor cylinder.
Replace the bacteriological filter
every 500 cycles or if dirty.
To change the filter:
1. Power the unit OFF,
disconnect tube A and remove
the filter from the bracket.
2. Disconnect tube B from the
back of the filter.
3. Before installing the replacement filter
(01-102119S) check that the arrow mark
on the filter matches the direction of
the arrow on the bracket. (Arrow should
be pointing to the right).
4. Insert filter into bracket and connect
tube B. Re-connect tube A.
5. After replacing the filter, go to the “Reset
LCS Counter” screen in the Setup menu
and reset the counter to 0.
Replace Water Filter
Replace Air Filter
CF 12 Water
Filter Clogged
CF 18 Water Filter
Pressure Sensor Stuck
CF 26 Air Valve error
CF 30 LCS adaptor
disconnected
Water filter in the LCS adaptor
is dirty or clogged.
The bacteriological air filter in the
kickplate is dirty.
Problem with Lumens Cleaning System.
Problem with Lumens Cleaning System.
Problem with air suppy pressure in air line.
Incoming air pressure must be between 1 bar
(15 psi) - 10 bar (145 psi).
Problem with Lumens Cleaning System.
Follow instructions above:
“Cleaning the water filter.”
Follow instructions above:
“Replacing the air filter.”
Check LCS adaptor is correctly positioned.
You selected LCS at program start but
LCS is not installed.
Replace Water Filter, see
“Cleaning the water filter.”
If cleaning doesn’t resolve the problem,
replace the water filter (01-113742S)
If message persists, call for service.
Call for service.
Call for service.
Check that LCS adaptor is correctly positioned.
Tube A Tube B (behind)
1 2 43
English
Adaptateur LCS et support
(compris)
Élément enfichable
LCS (compris)
Support pour chargement vertical
d’instruments (en option)
Système de nettoyage Lumens (LCS) pour modèles
HYDRIM
®
C61wd G4
Connexion des pièces à main à l’adaptateur LCS
Utilisation du menu d’écran tactile
Visser le LCS allant
avec le couplage
de pièce à main
approprié.
Introduire le couplage de pièce à main
dans celle-ci.
Pousser la pièce à main
dans l’une des fixations.
L’HYDRIM va détecter
automatiquement la
présence de l’adaptateur
LCS. Sélectionner le
programme désiré et
suivre les instructions.
Une fois terminé, ouvrir la
porte et retirer les pièces
à main de l’adaptateur en
poussant sur l’attache sur
le côté de la fixation.
Exemple : cycle régulier
1 2 3
L’adaptateur pour le système
de nettoyage Lumens (LCS)
ne pourra être utilisé qu’avec
les modèles HYDRIM C61wd
G4 équipés d’un système de
nettoyage Lumens. Si votre
HYDRIM peut travailler avec le
LCS, vous apercevrez un élément
enfichable blanc à l’intérieur côté
arrière gauche de la chambre.
1 2 3
French
Multiflex
Intramatic
Appuyer
sur l’attache
afin de
libérer.
Mise en place de l’adaptateur LCS
Retirer la vis ainsi que
l’élément enfichable
blanc de la paroi ar-
rière de la chambre de
l’appareil.
Retirez la bague de
raccord rapide, montez
l’accouplement du tuyau
d’air et libérez la bague
afin de verrouiller.
Fixer l’élément enfichable
blanc LCS sur le chariot
HYDRIM (lequel reste en
position basse lorsque le
support LCS n’est pas utilisé).
Fixer l’adaptateur LCS et le support
sur le chariot en engageant l’attache
en haut de l’élément enfichable LCS.
Placer le panier de nettoyage pour
canules sur les rails du support LCS.
1 2 43
Remarque : Conformément aux
recommandations des principaux
fabricants/fournisseurs de pièces
à main, les parties creuses des
instruments doivent être séchées
immédiatement après le nettoyage
et la désinfection en utilisant de l’air
comprimé médical avant d’effectuer
des procédés supplémentaires ou
une stérilisation. Cela doit être suivi
par les procédures d’entretien régulier
recommandées et par l’application
d’agents/huiles d’entretien. Veuillez
vous conformer à tous les règlements
en matière de santé et sécurité en
vigueur dans votre région.
Maintenance et dépannage
Messages système LCS Ce que cela signifie Que faire pour l’utilisateur?
NETTOYAGE DU FILTRE À EAU DU LCS
REMPLACEMENT DU FILTRE A AIR BACTÉRIOLOGIQUE
Remettre le filtre
dans le boîtier et bien
refermer le capuchon.
Enlever le filtre. Rincer le filtre et, si nécessaire,
nettoyer avec une brosse douce ou
à l’air comprimé. Vérifier et nettoyer
le joint torique. Enlever tous les
débris à l’intérieur du boîtier.
Dévisser le capuchon sur
la face avant du cylindre
d’adaptateur LCS.
Remplacer le filtre à air
bactériologique tous les 500 cycles
ou en cas d’encrassement.
Afin de changer le filtre :
1. Éteindre l’appareil, désolidariser
le tube A et enlever le filtre du
support.
2. Désolidariser le tube B de
l’arrière du filtre.
3. Avant d’installer le filtre de remplacement
(01-102119S), vérifier que la flèche de
marquage sur le filtre corresponde bien au
sens de celle du support (la flèche devant
pointer à droite).
4. Introduire le filtre dans le support et connecter
le tube B. Reconnecter ensuite le tube A.
5. Après avoir remplacé le filtre, aller à
l’écran « Réinitialiser compteur LCS »
au menu d’installation et réinitialiser le
compteur sur zéro.
Remplacer filtre à eau
Remplacer filtre à air
Filtre à eau
CF 12 bouché
Capteur de pression
filtre à eau CF 18 bloqué
Erreur vanne d’air CF 26
CF30 LCS
Adaptateur déconnecté
Le filtre à air de l’adaptateur LCS
est encrassé ou bouché.
Le filtre à air bactériologique dans
la plaque de pied est encrassé.
Problèmes causés par le système
de nettoyage des Lumens.
Problèmes causés par le système
de nettoyage des Lumens.
Problème causé par la pression dans
conduite d’alimentation en air.
L'entreé de la pression d’alimentation en air doit
être comprise entre 1 bar (15 psi)-10 bar (145 psi).
Problèmes causés par le système
de nettoyage des Lumens.
Suivre les instructions ci-dessus :
« Nettoyage du filtre à air »
Suivre les instructions ci-dessus :
« Remplacement du filtre à air »
Vérifier que l’adaptateur LCS soit
correctement positionné.
Vous avez sélectionné LCS lors du démarrage
du programme alors que LCS n’est pas installé.
Remplacer le filtre à eau, voir à cet effet
« Nettoyage du filtre à eau ».
Si le nettoyage n’est pas en mesure de résoudre
le problème, remplacer le filtre à eau (01-113742S).
Si le message persiste, appeler le service
de dépannage.
Appeler le service de dépannage.
Appeler le service de dépannage.
Vérifier que l’adaptateur LCS soit
correctement positionné.
Tube A Tube B (derrière)
1 2 43
French
LCS-Adapter und -Gestell
(inbegriffen)
LCS-Stopfen
(inbegriffen)
Racks fur vertikale Instrumente
(optional)
HYDRIM
®
C61wd G4 Innenreinigungssystem (LCS)
Anschließen der Handstücke an den LCS-Adapter
Eingaben über den Touchscreen
Schrauben Sie den
jeweiligen Handstück
anschluss auf die
LCS-Fassung.
Stecken Sie den Handstückanschluss in das
Handstück.
Drücken Sie das Handstück in
einen der Anschlussplätze.
Der HYDRIM erkennt den
LCS-Adapter automatisch.
Wählen Sie das gewünschte
Programm aus und befolgen
Sie die Anweisungen. Öffnen
Sie nach Programmende
die Tür und nehmen Sie das
Handstück vom Adapter
ab, indem Sie auf die
Lasche an der Seite des
Anschlussplatzes drücken.
Beispiel: Normaler Zyklus
1 2 3
Der Adapter für das
Innenreinigungssystem (LCS)
kann ausschließlich zusammen
mit HYDRIM-C61wd-G4-
Modellen verwendet werden,
die für die Verwendung eines
Innenreinigungssystems
vorgesehen sind. Ob Ihr
HYDRIM LCS-fähig ist, erkennen
Sie an einem weißen Stopfen
hinten links in der Waschkammer.
1 2 3
German
Multiflex
Intramatic
Drücken Sie
zum Lösen
die Lasche.
Montage des LCS-Adapters
Entfernen Sie die
Schraube und ziehen
Sie den weißen Stopfen
auf der Rückwand der
Kammer des Geräts heraus.
Ziehen Sie den
Schnellverschlussring
zurück, schieben Sie den
Druckluftstecker in die
Kupplung und lassen Sie
den Ring wieder los.
Befestigen Sie den LCS-
Stopfen am HYDRIM-
Rollwagen. (Bleibt in der
nach unten gerichteten
Position, wenn das LCS-
Gestell nicht in Gebrauch ist.)
Befestigen Sie den LCS-Adapter und
das Gestell am Rollwagen fixieren
Sie dabei die Lasche am oberen
Ende des LCS-Stopfens. Setzen Sie
den Kanülen-Reinigungskorb auf die
Schienen des LCS-Gestells.
1 2 43
Bitte beachten: Laut Empfehlungen
führender Hersteller/Lieferanten
von Handstücken sollten die hohlen
Instrumententeile sofort nach der
Reinigung und Desinfektion mit
medizinischer Druckluft getrocknet
werden, bevor weitere Verfahren
erfolgen oder eine Sterilisierung
vorgenommen wird. Im Anschluss
sind die empfohlenen normalen
Wartungsverfahren durchzuführen
und Pflegemittel/Öle aufzutragen.
Bitte befolgen Sie alle für Ihre Region
maßgeblichen Gesundheits- und
Sicherheitsvorschriften.
Instandhaltung und Fehlerbehebung
Meldungen des LCS-Systems Was die Meldung bedeutet Was zu tun ist – Benutzer
REINIGEN DES LCS-WASSERFILTERS
WECHSELN DES BAKTERIOLOGISCHEN LUFTFILTERS
Schieben Sie den Filter
wieder in das Gehäuse
und drehen Sie die
Kappe fest zu.
Nehmen Sie den Filter
heraus.
Spülen Sie den Filter aus und reinigen Sie
ihn – falls erforderlich – mit einer weichen
Bürste oder mit Hilfe von Druckluft.
Überprüfen und reinigen Sie den O-Ring.
Entfernen Sie jegliche Verunreinigungen
aus dem Inneren des Gehäuses.
Schrauben Sie die Kappe
auf der Vorderseite des
LCS-Adapterzylinders ab.
Wechseln Sie den
bakteriologischen Filter alle
500 Zyklen beziehungsweise
wenn der Filter verschmutzt ist.
So wechseln Sie den Filter:
1. Schalten Sie das Gerät ab
(OFF). Klemmen Sie Röhrchen
A ab und entnehmen Sie den
Filter aus der Halterung.
2. Klemmen Sie Röhrchen B von
der Rückseite des Filters ab.
3. Überprüfen Sie, bevor Sie den Austauschfilter
(01-102119S) installieren, ob der Pfeil auf dem
Filter in die Richtung des Pfeils auf der Halterung
zeigt. (Der Pfeil sollte nach rechts zeigen.)
4. Setzen Sie den Filter in die Halterung ein und
schließen Sie Röhrchen B an. Schließen Sie
Röhrchen A wieder an.
5. Nachdem Sie den Filter ausgetauscht haben,
gehen Sie im Setup-Menü in den Bildschirm
„LCS-Zähler zurücksetzen“ und setzen den
Zähler auf null zurück.
Wasserfilter austauschen
Luftfilter austauschen
CF 12 Wasserfilter verschmutzt
CF 18 Drucksensor des
Wasserfilters blockiert
CF 26 Fehler Luftventil
CF 30 LCS-Adapter getrennt
Wasserfilter im LCS-Adapter ist
verschmutzt oder zugesetzt.
Der bakteriologische Luftfilter im
Sockel-Frontblech ist verschmutzt.
Problem mit dem
Innenreinigungssystem.
Problem mit dem
Innenreinigungssystem.
Fehler in Druckluftzufuhr.
Eingehende Luftzufuhr muss zwischen
1 bar (15 psi)-10 bar (145 psi) sein
Problem mit dem
Innenreinigungssystem.
Befolgen Sie die oben stehenden Anweisungen
unter: „Reinigen des Wasserfilters LCS “
Befolgen Sie die oben stehenden Anweisungen
unter: „Wechseln des bakteriologischen Luftfilters“
Überprüfen Sie, ob der LCS-Adapter
ordnungsgemäß angebracht ist.
Möglicherweise haben Sie bei Programmbeginn „LCS“
ausgewählt, obwohl kein LCS installiert ist.
Tauschen Sie den Wasserfilter aus, siehe
„Reinigen des Wasserfilters“.
Wenn das Problem durch Reinigen nicht behoben
wird, tauschen Sie den Wasserfilter (01-113742S) aus.
Wenn die Meldung weiterhin angezeigt wird, wenden
Sie sich an den technischen Service.
Wenden Sie sich an den technischen Service.
Wenden Sie sich an den technischen Service.
Überprüfen Sie, ob der LCS-Adapter
ordnungsgemäß angebracht ist.
Röhrchen A Röhrchen B (dahinter)
1 2 43
German
Adaptador y gradilla de SLL
(incluido)
Tapón de
SLL (incluido)
Sistema de limpieza de lúmenes (SLL)
en HYDRIM
®
C61wd G4
Conexión de piezas de mano al adaptador de SLL
Uso del menú de la pantalla táctil
Enrosque el conector
de SLL al
correspondiente
acoplamiento de
pieza de mano.
Inserte el acoplamiento en la pieza de mano. Introduzca la pieza de mano
en uno de los puertos.
La unidad HYDRIM detec-
tará automáticamente la
presencia del adaptador de
SLL. Seleccione el programa
deseado y siga las indica-
ciones. Una vez comple-
tados los pasos, abra la
puerta y extraiga las piezas
de mano del adaptador pre-
sionando la pestaña situada
al lado del puerto.
Ejemplo: Ciclo Normal
1 2 3
El adaptador para sistema de
limpieza de lúmenes (SLL) sólo
puede ser utilizado con los
modelos HYDRIM C61wd G4 que
se encuentran equipados con un
sistema de limpieza de lúmenes.
Si su unidad HYDRIM está
habilitada para SLL, usted verá un
tapón blanco en la parte trasera
interior izquierda de la cámara.
1 2 3
Spanish
Multiflex
Intramatic
Presione
la pestaña
para soltar
Gradilla para instrumentos
verticales (opcional)
Ajuste del adaptador de SLL
Quite el tornillo y extraiga
el tapón blanco situado
en la pared trasera de la
cámara de la unidad.
Tire hacia atrás el anillo
de montaje de conexión
rápida, inserte el
acoplamiento de la línea de
aire y suelte el anillo para
bloquear.
Fije el tapón de SLL
al carrito de la unidad
HYDRIM. (Se mantiene en
posición baja mientras
no se utiliza la gradilla de
SLL.)
Fije el adaptador y gradilla de SLL al
carrito, enganchando la pestaña en
la parte superior del tapón. Coloque
la cesta de limpieza de cánulas sobre
las guías de la gradilla de SLL.
1 2 43
Nota: como recomiendan los
principales fabricantes/proveedores
de piezas de mano, las partes
huecas de los instrumentos deben
secarse inmediatamente después de
limpiarlas y desinfectarlas con aire
comprimido médico antes de realizar
procesos y esterilización adicionales.
A continuación deben realizarse los
procedimientos de mantenimiento
habituales recomendados y la
aplicación de los agentes/aceites
de mantenimiento. Siga todas las
normativas sanitarias y de seguridad
para su región.
Mantenimiento y resolución de problemas
Mensajes del SLL Qué significa Qué hacer: Usuario
LIMPIEZA DEL FILTRO DE AGUA DE SLL
REEMPLAZO DEL FILTRO DE AIRE BACTERIOLÓGICO
Reinserte el filtro en la
carcasa y cierre bien
la tapa.
Extraiga el filtro. Enjuague el filtro y, en caso necesario,
limpie con un cepillo suave o con aire
comprimido. Controle y limpie la junta
tórica. Quite cualquier residuo que
haya quedado dentro de la carcasa.
Desenrosque la tapa
dispuesta en la parte
delantera del cilindro
del adaptador de SLL.
Reemplace el filtro
bacteriológico cada 500
ciclos o cuando está sucio.
Para cambiar el filtro:
1. Apague la unidad,
desconecte el tubo A y
quite el filtro del soporte.
2. Desconecte el tubo B de
la parte posterior del filtro.
3. Antes de instalar el filtro de reemplazo
(01-102119S), verifique que la marca de la
flecha del filtro coincida con la dirección de la
flecha situada en el soporte. (La flecha debe
apuntar hacia la derecha.)
4. Inserte el filtro en el soporte y conecte el tubo B.
Vuelva a conectar el tubo A.
5. Una vez reemplazado el filtro, vaya a la pantalla
“Restablecer contador de SLL” en el menú
Configuración y vuelva a colocar en cero el
contador.
Reemplazar filtro de agua
Reemplazar filtro de aire
CF 12 Filtro de agua
obstruido
CF 18 Sensor de presión
del filtro de agua atascado
CF 26 Error en válvula
de aire
CF 30 Adaptador
de SLL desconectado
El filtro de agua en el adaptador
de SLL está sucio u obstruido.
El filtro de aire bacteriológico
en la placa de protección está sucio.
Problema con el sistema de limpieza
de lúmenes.
Problema con el sistema de limpieza
de lúmenes.
Problema con la presión de alimentación
en la línea de aire.
La entrada de la presión de suministro de aire
debe ser de entre 1 bar (15 psi)-10 bar (145 psi)
Problema con el sistema de limpieza
de lúmenes.
Siga las instrucciones mencionadas
anteriormente: “Limpieza del filtro de agua”.
Siga las instrucciones mencionadas
anteriormente: “Reemplazo del filtro de aire”.
Compruebe que el adaptador de SLL
esté correctamente posicionado.
Usted seleccionó SLL al inicio del programa,
pero SLL no está instalado.
Reemplace el filtro de agua, ver
“Limpieza del filtro de agua”.
Si la limpieza no resuelve el problema,
reemplace el filtro de agua (01-113742S).
Si el mensaje sigue apareciendo, llame al
servicio técnico.
Llame al servicio técnico.
Llame al servicio técnico.
Compruebe que el adaptador de SLL esté
correctamente posicionado.
Tubo A Tubo B (detrás)
1 2 43
Spanish
Adattatore SPL e rack
(inclusi)
Spinotto SPL
(incluso)
Rack per strumenti verticali (optional)
HYDRIM
®
C61wd G4 Sistema di pulizia dei lumi (SPL)
Collegamento dei manipoli all’adattatore SPL
Utilizzo del menu touchscreen
Avvitare il raccordo
LCS al corrispondente
attacco per manipoli.
Inserire l’attacco per manipoli nel manipolo. Infilare il manipolo in una
delle prese.
L’apparecchio HYDRIM
rileverà automaticamente
la presenza dell’adattatore
SPL. Selezionare il
programma desiderato
e seguire le istruzioni. Al
termine della procedura,
aprire il portello e rimuovere
i manipoli dall’adattatore
spingendo la linguetta a lato
del portello.
Esempio: ciclo regolare
1 2 3
L’adattatore per il sistema di
pulizia dei lumi (SPL) può essere
utilizzato solo con i modelli
HYDRIM C61wd G4 dotati di un
sistema di pulizia dei lumi. Se
l’apparecchio HYDRIM è abilitato
per SPL, sulla parete posteriore
interna di sinistra della camera è
presente un tappo bianco.
1 2 3
Italian
Multiflex
Intramatic
Premere la
linguetta per
lo sgancio
Montaggio dell’adattatore SPL
Togliere la vite ed estrarre
il tappo bianco presente
sulla parete posteriore
della camera dell’unità.
Retrarre l’anello di
adattamento a innesto
rapido, inserire l’attacco
della linea pneumatica e
rilasciare l’anello in modo
che si blocchi.
Collegare il tappo SPL
al carrello HYDRIM. (è
in posizione abbassata
quando il rack SPL non
viene utilizzato)
Collegare l’adattatore SPL e il rack
al carrello, agganciando la linguetta
presente nella parte superiore dello
spinotto SPL. Posizionare il cestello di
pulizia cannule sui binari del rack SPL.
1 2 43
Notare che: come raccomandato dai
produttori/fornitori leader di manipoli,
le parti cave degli strumenti devono
essere asciugate immediatamente
dopo la pulizia e la disinfezione con
aria compressa per uso medico, prima
di effettuare ulteriori processi o la
sterilizzazione. Quindi, effettuare le
regolari procedure di manutenzione
raccomandate e applicare gli agenti/oli
di manutenzione. Si prega di seguire
tutti i regolamenti locali in vigore in
materia di salute e sicurezza.
Manutenzione e risoluzione dei problemi
Messaggi sistema LCS Significato Rimedio – Utente
PULIZIA DEL FILTRO ACQUA SPL
SOSTITUZIONE DEL FILTRO BATTERIOLOGICO
Reinserire il filtro
nell’alloggiamento e
chiudere saldamente
il tappo.
Togliere il filtro. Lavare il filtro e, se necessario, pulirlo
con una spazzola morbida o con
aria compressa. Controllare e pulire
l’O-ring. Togliere eventuali impurità
presenti all’interno dell’alloggiamento.
Svitare il tappo sul lato
anteriore del cilindro
adattatore SPL.
Sostituire il filtro batteriologico
ogni 500 cicli o se è sporco.
Per sostituire il filtro:
1. Spegnere l’unità, staccare
il tubo A e togliere il filtro
dalla staffa.
2. Scollegare il tubo B dalla
parte posteriore del filtro.
3. Prima di installare il filtro di ricambio
(01-102119S) verificare che il senso della
freccia sul filtro corrisponda al senso della
freccia sulla staffa. (La freccia dovrebbe
indicare a destra).
4. Inserire il filtro nella staffa e collegare il
tubo B. Ricollegare il tubo A.
5. Dopo aver sostituito il filtro, entrare nella
schermata “Resettare contat. SPL” nel
menu Setup e resettare il contatore a 0.
Sostituire filtro acqua
Sostituire filtro aria
CF 12 Filtro acqua intasato
CF 18 Sensore pressione
filtro acqua bloccato
CF 26 Errore valvola aria
CF 30 Adattatore
SPL scollegato
Il filtro acqua nell’adattatore SPL
è sporco o intasato.
Il filtro aria batteriologico nel
fermapiede è sporco.
Problema con il sistema di pulizia dei lumi.
Problema con il sistema di pulizia dei lumi.
Problema con la pressione di alimentazione
dell’aria nella linea pneumatica.
Incoming pressione di alimentazione di aria
deve essere compresa tra 1 bar (15 psi)-10 bar
(145 psi)
Problema con il sistema di pulizia dei lumi.
Seguire le istruzioni precedentemente
menzionate: “Pulizia del filtro acqua .”
Seguire le istruzioni precedentemente
menzionate: “Sostituzione del filtro dell’aria.”
Controllare se l’adattatore SPL è correttamente
posizionato.
È stato selezionato l’SPL all’avvio del
programma, ma l’SPL non è installato.
Sostituire il filtro acqua, vedere
“Pulizia del filtro acqua.”
Se la pulizia non risolve il problema,
sostituire il filtro acqua (01-113742S)
Se il messaggio persiste, rivolgersi
all’assistenza.
Rivolgersi all’assistenza.
Rivolgersi all’assistenza.
Controllare che l’adattatore SPL sia
correttamente posizionato.
Tubo A Tubo B (dietro)
1 2 43
Italian
LCS 插头和网架
(包括)
LCS 插头
(包括)
垂向器械网架(可选)
HYDRIM
®
C61wd G4 内腔清洗系统 (LCS)
将手机与 LCS 插头连接
使用触摸屏菜单
拧紧 LCS,安装到合适
的手机接头。
将手机接头插入手机。 将手机推入一个端口中。
HYDRIM 将会自动检测
LCS 插头的存在。 选择
所需的程序,并按照提示
操作。 完成后,打开设
备门,通过在端口侧面推
按选项取出手机。
举例: 常规循环
1 2 3
内腔清洗系统 (LCS) 接头只能用
于配备内腔清洗系统的 HYDRIM
C61wd G4 机型。 如果您的
HYDRIM能够配合LCS使用,您
将会在腔内左后方上看到白色插
头。
1 2 3
Chinese
Multiflex
Intramatic
按下选项以
释放
设置 LCS 插头
在设备腔后壁上拆下
螺丝并拔出白色插头
回拉快速接头连接环,插
入空气管道接头并释放
环,以便锁定。
LCS 插头连接到
HYDRIM 滚轮。 (未使用
LCS 网架时保持向下位置)
LCS 插头和网架连接到滚轮,在
LCS 插头的顶部占用选项。 将套管
清洗篮置于 LCS 网架的栏杆上。
1 2 43
请注意:根据领先的机头制造
商/供应商的建议,仪器中空
部分在清洗和消毒之后应该立
即使用医用压缩空气进行干
燥,然后才能进行额外的处理
或灭菌。在这之后,应该执行
推荐的例行维护程序,并施加
保养剂/保养油。请遵循您所
在地区所有适用的卫生与安全
监管规定。
保养与故障排除
LCS 系统消息 这意味着什么 怎么办 - 用户
清洁 LCS 水过滤器
更换空气过滤器
将过滤器重新装到壳体
上并紧紧关闭盖子。
取下过滤器。 冲洗过滤器,如果需要,用软刷或用压
缩空气进行清洁。 检查并清洁 O 型密
封圈。 清除壳体内部的任何残余物。
LCS 插头圆筒的前方
旋开盖子。
500 个循环或如果已脏,
则更换空气过滤器。
要更换过滤器:
1. 关闭设备电源并断开 A 管,
然后从托架上取出过滤器。
2. 从过滤器的后面断开 B 管。
3. 在安装替代的过滤器 (01-102119S) 之前,
请检查过滤器机器上的箭头标记与托架上
的箭头方向相符。 (箭头应当指向右侧)。
4. 将过滤器插入托架并连接 B 管。重新连接
A 管。
5. 在更换过滤器后,转到设置菜单中的重置
LCS 计数器屏幕并且将计数器重置为 0
更换水过滤器
更换空气过滤器
CF 12
过滤器阻塞
CF 18 水过滤器
压力传感器卡住
CF 26 空气阀错误
CF 30 LCS 插头 断开
LCS 中的水过滤器已脏或者堵塞。
踢板中的空气过滤器已脏。
内腔清洗系统问题。
内腔清洗系统问题。
空气管道中供气压力问题。
空气应压力必1 bar (15 psi)至
10 bar (145 psi)之间。
内腔清洗系统问题。
遵照上面的说明:清洁 LCS 水过滤器。
遵照上面的说明: “更换细菌空气过滤器。”
检查 LCS 插头是否正确定位。
您在程序启动处选择了 LCS 但是 LCS 未安装。
更换水过滤器,参见 “清洁 LCS 水过滤器。”
如果清洁并未解决问题,请更换水过滤器
(01-113742S)
如信息仍存在,请致电服务部。
请致电经销商。
请致电经销商。
检查 LCS 插头是否正确定位。
A B 管(后面)
1 2 43
Chinese
www.scican.com
Manufactured by
SciCan Ltd.
1440 Don Mills Road
Toronto, Ontario M3B 3P9
Tel: 416-445-1600
Toll free: 1-800-667-7733
Fax: 416-445-2727
EU Rep:
SciCan GmbH
Wangener Strasse 78
88299 Leutkirch, GERMANY
Tel.: +49 (0)7561 98343 - 0
Fax: +49 (0)7561 98343 - 699
SciCan Medtech AG
Alpenstrasse 16
6300 Zug, SWITZERLAND
Tel.: +41 (0) 41-727 7027
Fax.: +41 (0) 41-727 7029
0123
HYDRIM is a registered trademark and Your Infection Control Specialist is a trademark of SciCan Ltd. 95-113906 EU EFIGSC R2
SciCan
@SciCan
SciCanTV

Transcripción de documentos

• Operator’s Manual • Mode d’emploi • Bedienerhandbuch • Manuale d’uso • Manual del operador • 操作手册 95-113906 EU ML R4. HYDRIM C61wd G4 Lumens Cleaning System (LCS) Installation Guide and Operator’s Manual Copyright 2015 SciCan Ltd. All rights reserved. ® HYDRIM C61wd G4 LUMENS CLEANING SYSTEM (LCS) English HYDRIM ® C61wd G4 Lumens Cleaning System (LCS) The Lumens Cleaning System (LCS) Adaptor can only be used with HYDRIM C61wd G4 models equipped with a lumens cleaning system. If your HYDRIM is LCS enabled, you will see a white plug on the inside back left of the chamber. Rack for vertical instruments (optional) LCS Plug (included) LCS Adaptor and Rack (included) Setting up the LCS unit 1 2 Pull the quick connect fitting ring back, insert air line coupling and release the ring to lock. 3 Remove the screw and pull out the white plug on the back wall of the unit’s chamber. 4 Attach the LCS plug to the HYDRIM trolley. (Stays in the down position when LCS rack is not being used) Attach the LCS adaptor and rack to the trolley, engaging the tab at the top of the LCS plug. Place the cannula cleaning basket on the rails of the LCS rack. Connecting handpieces to the LCS adaptor 1 2 3 Insert the handpiece coupling into the handpiece. Push the handpiece into one of the ports. Push tab to release Multiflex Screw the LCS fitting to the appropriate handpiece coupling. Intramatic Using the touchscreen menu The HYDRIM will automatically detect the presence of the LCS adaptor. Select the desired program and follow the prompts. Upon completion, open the door and remove the handpieces from the adaptor by pushing on the tab at the side of the port. Example: Regular cycle 1 2 3 Please note: As recommended by leading manufacturer/suppliers of handpieces, the hollow parts of instruments should be dried immediately after cleaning and disinfection using medical compressed air before additional processes or sterilization. This is to be followed by recommended regular maintenance procedures and application of maintenance agents/oils. Please follow all applicable health & safety regulations for your region. English Maintenance and troubleshooting CLEANING THE LCS WATER FILTER 1 2 3 4 Unscrew the cap at the front of the LCS adaptor cylinder. Remove the filter. Rinse the filter and, if required, clean with a soft brush or using compressed air. Check and clean the O-ring. Remove any debris inside the housing. Return the filter to the housing and close the cap tightly. REPLACING THE BACTERIOLOGICAL AIR FILTER Replace the bacteriological filter every 500 cycles or if dirty. To change the filter: 1. Power the unit OFF, disconnect tube A and remove the filter from the bracket. 2. Disconnect tube B from the back of the filter. Tube A Tube B (behind) 3. Before installing the replacement filter (01-102119S) check that the arrow mark on the filter matches the direction of the arrow on the bracket. (Arrow should be pointing to the right). 4. Insert filter into bracket and connect tube B. Re-connect tube A. 5. After replacing the filter, go to the “Reset LCS Counter” screen in the Setup menu and reset the counter to 0. LCS system Messages What it Means What to Do – User Replace Water Filter Water filter in the LCS adaptor is dirty or clogged. • Follow instructions above: “Cleaning the water filter.” Replace Air Filter The bacteriological air filter in the kickplate is dirty. • Follow instructions above: “Replacing the air filter.” • Check LCS adaptor is correctly positioned. • You selected LCS at program start but LCS is not installed. CF 12 Water Filter Clogged Problem with Lumens Cleaning System. • Replace Water Filter, see “Cleaning the water filter.” • If cleaning doesn’t resolve the problem, replace the water filter (01-113742S) • If message persists, call for service. CF 18 Water Filter Pressure Sensor Stuck Problem with Lumens Cleaning System. • Call for service. CF 26 Air Valve error Problem with air suppy pressure in air line. Incoming air pressure must be between 1 bar (15 psi) - 10 bar (145 psi). • Call for service. CF 30 LCS adaptor disconnected Problem with Lumens Cleaning System. • Check that LCS adaptor is correctly positioned. French Système de nettoyage Lumens (LCS) pour modèles HYDRIM ® C61wd G4 L’adaptateur pour le système de nettoyage Lumens (LCS) ne pourra être utilisé qu’avec les modèles HYDRIM C61wd G4 équipés d’un système de nettoyage Lumens. Si votre HYDRIM peut travailler avec le LCS, vous apercevrez un élément enfichable blanc à l’intérieur côté arrière gauche de la chambre. Support pour chargement vertical d’instruments (en option) Élément enfichable LCS (compris) Adaptateur LCS et support (compris) Mise en place de l’adaptateur LCS 1 Retirez la bague de raccord rapide, montez l’accouplement du tuyau d’air et libérez la bague afin de verrouiller. 2 3 4 Fixer l’élément enfichable blanc LCS sur le chariot HYDRIM (lequel reste en position basse lorsque le support LCS n’est pas utilisé). Retirer la vis ainsi que l’élément enfichable blanc de la paroi arrière de la chambre de l’appareil. Fixer l’adaptateur LCS et le support sur le chariot en engageant l’attache en haut de l’élément enfichable LCS. Placer le panier de nettoyage pour canules sur les rails du support LCS. Connexion des pièces à main à l’adaptateur LCS 1 Appuyer sur l’attache afin de libérer. 2 3 Introduire le couplage de pièce à main dans celle-ci. Pousser la pièce à main dans l’une des fixations. Multiflex Visser le LCS allant avec le couplage de pièce à main approprié. Intramatic Utilisation du menu d’écran tactile L’HYDRIM va détecter automatiquement la présence de l’adaptateur LCS. Sélectionner le programme désiré et suivre les instructions. Une fois terminé, ouvrir la porte et retirer les pièces à main de l’adaptateur en poussant sur l’attache sur le côté de la fixation. Exemple : cycle régulier 1 2 3 Remarque : Conformément aux recommandations des principaux fabricants/fournisseurs de pièces à main, les parties creuses des instruments doivent être séchées immédiatement après le nettoyage et la désinfection en utilisant de l’air comprimé médical avant d’effectuer des procédés supplémentaires ou une stérilisation. Cela doit être suivi par les procédures d’entretien régulier recommandées et par l’application d’agents/huiles d’entretien. Veuillez vous conformer à tous les règlements en matière de santé et sécurité en vigueur dans votre région. French Maintenance et dépannage NETTOYAGE DU FILTRE À EAU DU LCS 1 2 3 4 Dévisser le capuchon sur la face avant du cylindre d’adaptateur LCS. Enlever le filtre. Rincer le filtre et, si nécessaire, nettoyer avec une brosse douce ou à l’air comprimé. Vérifier et nettoyer le joint torique. Enlever tous les débris à l’intérieur du boîtier. Remettre le filtre dans le boîtier et bien refermer le capuchon. REMPLACEMENT DU FILTRE A AIR BACTÉRIOLOGIQUE Remplacer le filtre à air bactériologique tous les 500 cycles ou en cas d’encrassement. Afin de changer le filtre : 1. Éteindre l’appareil, désolidariser le tube A et enlever le filtre du support. 2. Désolidariser le tube B de l’arrière du filtre. Tube A Tube B (derrière) 3. Avant d’installer le filtre de remplacement (01-102119S), vérifier que la flèche de marquage sur le filtre corresponde bien au sens de celle du support (la flèche devant pointer à droite). 4. Introduire le filtre dans le support et connecter le tube B. Reconnecter ensuite le tube A. 5. Après avoir remplacé le filtre, aller à l’écran « Réinitialiser compteur LCS » au menu d’installation et réinitialiser le compteur sur zéro. Messages système LCS Ce que cela signifie Que faire pour l’utilisateur? Remplacer filtre à eau Le filtre à air de l’adaptateur LCS est encrassé ou bouché. • Suivre les instructions ci-dessus : « Nettoyage du filtre à air » Remplacer filtre à air Le filtre à air bactériologique dans la plaque de pied est encrassé. • Suivre les instructions ci-dessus : « Remplacement du filtre à air » • Vérifier que l’adaptateur LCS soit correctement positionné. • Vous avez sélectionné LCS lors du démarrage du programme alors que LCS n’est pas installé. Filtre à eau CF 12 bouché Problèmes causés par le système de nettoyage des Lumens. • Remplacer le filtre à eau, voir à cet effet « Nettoyage du filtre à eau ». • Si le nettoyage n’est pas en mesure de résoudre le problème, remplacer le filtre à eau (01-113742S). • Si le message persiste, appeler le service de dépannage. Capteur de pression filtre à eau CF 18 bloqué Problèmes causés par le système de nettoyage des Lumens. • Appeler le service de dépannage. Erreur vanne d’air CF 26 Problème causé par la pression dans conduite d’alimentation en air. • Appeler le service de dépannage. L'entreé de la pression d’alimentation en air doit être comprise entre 1 bar (15 psi)-10 bar (145 psi). CF30 LCS Adaptateur déconnecté Problèmes causés par le système de nettoyage des Lumens. • Vérifier que l’adaptateur LCS soit correctement positionné. German HYDRIM ® C61wd G4 Innenreinigungssystem (LCS) Der Adapter für das Innenreinigungssystem (LCS) kann ausschließlich zusammen mit HYDRIM-C61wd-G4Modellen verwendet werden, die für die Verwendung eines Innenreinigungssystems vorgesehen sind. Ob Ihr HYDRIM LCS-fähig ist, erkennen Sie an einem weißen Stopfen hinten links in der Waschkammer. Racks fur vertikale Instrumente (optional) LCS-Stopfen (inbegriffen) LCS-Adapter und -Gestell (inbegriffen) Montage des LCS-Adapters 1 Ziehen Sie den Schnellverschlussring zurück, schieben Sie den Druckluftstecker in die Kupplung und lassen Sie den Ring wieder los. 2 3 Entfernen Sie die Schraube und ziehen Sie den weißen Stopfen auf der Rückwand der Kammer des Geräts heraus. 4 Befestigen Sie den LCSStopfen am HYDRIMRollwagen. (Bleibt in der nach unten gerichteten Position, wenn das LCSGestell nicht in Gebrauch ist.) Befestigen Sie den LCS-Adapter und das Gestell am Rollwagen fixieren Sie dabei die Lasche am oberen Ende des LCS-Stopfens. Setzen Sie den Kanülen-Reinigungskorb auf die Schienen des LCS-Gestells. Anschließen der Handstücke an den LCS-Adapter 1 2 3 Stecken Sie den Handstückanschluss in das Handstück. Drücken Sie das Handstück in einen der Anschlussplätze. Drücken Sie zum Lösen die Lasche. Multiflex Schrauben Sie den jeweiligen Handstück anschluss auf die LCS-Fassung. Intramatic Eingaben über den Touchscreen Der HYDRIM erkennt den LCS-Adapter automatisch. Wählen Sie das gewünschte Programm aus und befolgen Sie die Anweisungen. Öffnen Sie nach Programmende die Tür und nehmen Sie das Handstück vom Adapter ab, indem Sie auf die Lasche an der Seite des Anschlussplatzes drücken. Beispiel: Normaler Zyklus 1 2 3 Bitte beachten: Laut Empfehlungen führender Hersteller/Lieferanten von Handstücken sollten die hohlen Instrumententeile sofort nach der Reinigung und Desinfektion mit medizinischer Druckluft getrocknet werden, bevor weitere Verfahren erfolgen oder eine Sterilisierung vorgenommen wird. Im Anschluss sind die empfohlenen normalen Wartungsverfahren durchzuführen und Pflegemittel/Öle aufzutragen. Bitte befolgen Sie alle für Ihre Region maßgeblichen Gesundheits- und Sicherheitsvorschriften. German Instandhaltung und Fehlerbehebung REINIGEN DES LCS-WASSERFILTERS 1 2 3 4 Schrauben Sie die Kappe auf der Vorderseite des LCS-Adapterzylinders ab. Nehmen Sie den Filter heraus. Spülen Sie den Filter aus und reinigen Sie ihn – falls erforderlich – mit einer weichen Bürste oder mit Hilfe von Druckluft. Überprüfen und reinigen Sie den O-Ring. Entfernen Sie jegliche Verunreinigungen aus dem Inneren des Gehäuses. Schieben Sie den Filter wieder in das Gehäuse und drehen Sie die Kappe fest zu. WECHSELN DES BAKTERIOLOGISCHEN LUFTFILTERS Wechseln Sie den bakteriologischen Filter alle 500 Zyklen beziehungsweise wenn der Filter verschmutzt ist. 3. Überprüfen Sie, bevor Sie den Austauschfilter (01-102119S) installieren, ob der Pfeil auf dem Filter in die Richtung des Pfeils auf der Halterung zeigt. (Der Pfeil sollte nach rechts zeigen.) So wechseln Sie den Filter: 4. Setzen Sie den Filter in die Halterung ein und schließen Sie Röhrchen B an. Schließen Sie Röhrchen A wieder an. 1. S  chalten Sie das Gerät ab (OFF). Klemmen Sie Röhrchen A ab und entnehmen Sie den Filter aus der Halterung. Röhrchen A Röhrchen B (dahinter) 2. Klemmen Sie Röhrchen B von der Rückseite des Filters ab. 5. Nachdem Sie den Filter ausgetauscht haben, gehen Sie im Setup-Menü in den Bildschirm „LCS-Zähler zurücksetzen“ und setzen den Zähler auf null zurück. Meldungen des LCS-Systems Was die Meldung bedeutet Was zu tun ist – Benutzer Wasserfilter austauschen Wasserfilter im LCS-Adapter ist verschmutzt oder zugesetzt. • Befolgen Sie die oben stehenden Anweisungen unter: „Reinigen des Wasserfilters LCS “ Luftfilter austauschen Der bakteriologische Luftfilter im Sockel-Frontblech ist verschmutzt. • Befolgen Sie die oben stehenden Anweisungen unter: „Wechseln des bakteriologischen Luftfilters“ • Überprüfen Sie, ob der LCS-Adapter ordnungsgemäß angebracht ist. • Möglicherweise haben Sie bei Programmbeginn „LCS“ ausgewählt, obwohl kein LCS installiert ist. CF 12 Wasserfilter verschmutzt Problem mit dem Innenreinigungssystem. • Tauschen Sie den Wasserfilter aus, siehe „Reinigen des Wasserfilters“. • Wenn das Problem durch Reinigen nicht behoben wird, tauschen Sie den Wasserfilter (01-113742S) aus. • Wenn die Meldung weiterhin angezeigt wird, wenden Sie sich an den technischen Service.  F 18 Drucksensor des C Wasserfilters blockiert Problem mit dem Innenreinigungssystem. • Wenden Sie sich an den technischen Service. CF 26 Fehler Luftventil Fehler in Druckluftzufuhr. Eingehende Luftzufuhr muss zwischen 1 bar (15 psi)-10 bar (145 psi) sein • Wenden Sie sich an den technischen Service. CF 30 LCS-Adapter getrennt Problem mit dem Innenreinigungssystem. • Überprüfen Sie, ob der LCS-Adapter ordnungsgemäß angebracht ist. Spanish Sistema de limpieza de lúmenes (SLL) en HYDRIM ® C61wd G4 El adaptador para sistema de limpieza de lúmenes (SLL) sólo puede ser utilizado con los modelos HYDRIM C61wd G4 que se encuentran equipados con un sistema de limpieza de lúmenes. Si su unidad HYDRIM está habilitada para SLL, usted verá un tapón blanco en la parte trasera interior izquierda de la cámara. Gradilla para instrumentos verticales (opcional) Tapón de SLL (incluido) Adaptador y gradilla de SLL (incluido) Ajuste del adaptador de SLL 1 Tire hacia atrás el anillo de montaje de conexión rápida, inserte el acoplamiento de la línea de aire y suelte el anillo para bloquear. 2 3 Quite el tornillo y extraiga el tapón blanco situado en la pared trasera de la cámara de la unidad. 4 Fije el tapón de SLL al carrito de la unidad HYDRIM. (Se mantiene en posición baja mientras no se utiliza la gradilla de SLL.) Fije el adaptador y gradilla de SLL al carrito, enganchando la pestaña en la parte superior del tapón. Coloque la cesta de limpieza de cánulas sobre las guías de la gradilla de SLL. Conexión de piezas de mano al adaptador de SLL 1 2 3 Inserte el acoplamiento en la pieza de mano. Introduzca la pieza de mano en uno de los puertos. Presione la pestaña para soltar Multiflex Enrosque el conector de SLL al correspondiente acoplamiento de pieza de mano. Intramatic Uso del menú de la pantalla táctil La unidad HYDRIM detectará automáticamente la presencia del adaptador de SLL. Seleccione el programa deseado y siga las indicaciones. Una vez completados los pasos, abra la puerta y extraiga las piezas de mano del adaptador presionando la pestaña situada al lado del puerto. Ejemplo: Ciclo Normal 1 2 3 Nota: como recomiendan los principales fabricantes/proveedores de piezas de mano, las partes huecas de los instrumentos deben secarse inmediatamente después de limpiarlas y desinfectarlas con aire comprimido médico antes de realizar procesos y esterilización adicionales. A continuación deben realizarse los procedimientos de mantenimiento habituales recomendados y la aplicación de los agentes/aceites de mantenimiento. Siga todas las normativas sanitarias y de seguridad para su región. Spanish Mantenimiento y resolución de problemas LIMPIEZA DEL FILTRO DE AGUA DE SLL 1 2 3 4 Desenrosque la tapa dispuesta en la parte delantera del cilindro del adaptador de SLL. Extraiga el filtro. Enjuague el filtro y, en caso necesario, limpie con un cepillo suave o con aire comprimido. Controle y limpie la junta tórica. Quite cualquier residuo que haya quedado dentro de la carcasa. Reinserte el filtro en la carcasa y cierre bien la tapa. REEMPLAZO DEL FILTRO DE AIRE BACTERIOLÓGICO Reemplace el filtro bacteriológico cada 500 ciclos o cuando está sucio. Para cambiar el filtro: 1. Apague la unidad, desconecte el tubo A y quite el filtro del soporte. 2. Desconecte el tubo B de la parte posterior del filtro. Tubo A Tubo B (detrás) 3. Antes de instalar el filtro de reemplazo (01-102119S), verifique que la marca de la flecha del filtro coincida con la dirección de la flecha situada en el soporte. (La flecha debe apuntar hacia la derecha.) 4. Inserte el filtro en el soporte y conecte el tubo B. Vuelva a conectar el tubo A. 5. Una vez reemplazado el filtro, vaya a la pantalla “Restablecer contador de SLL” en el menú Configuración y vuelva a colocar en cero el contador. Mensajes del SLL Qué significa Qué hacer: Usuario Reemplazar filtro de agua El filtro de agua en el adaptador de SLL está sucio u obstruido. • Siga las instrucciones mencionadas anteriormente: “Limpieza del filtro de agua”. Reemplazar filtro de aire El filtro de aire bacteriológico en la placa de protección está sucio. • Siga las instrucciones mencionadas anteriormente: “Reemplazo del filtro de aire”. • Compruebe que el adaptador de SLL esté correctamente posicionado. • Usted seleccionó SLL al inicio del programa, pero SLL no está instalado. CF 12 Filtro de agua obstruido Problema con el sistema de limpieza de lúmenes. • Reemplace el filtro de agua, ver “Limpieza del filtro de agua”. • Si la limpieza no resuelve el problema, reemplace el filtro de agua (01-113742S). • Si el mensaje sigue apareciendo, llame al servicio técnico. CF 18 Sensor de presión del filtro de agua atascado Problema con el sistema de limpieza de lúmenes. • Llame al servicio técnico. CF 26 Error en válvula de aire Problema con la presión de alimentación en la línea de aire. La entrada de la presión de suministro de aire debe ser de entre 1 bar (15 psi)-10 bar (145 psi) • Llame al servicio técnico. CF 30 Adaptador de SLL desconectado Problema con el sistema de limpieza de lúmenes. • Compruebe que el adaptador de SLL esté correctamente posicionado. Italian HYDRIM ® C61wd G4 Sistema di pulizia dei lumi (SPL) L’adattatore per il sistema di pulizia dei lumi (SPL) può essere utilizzato solo con i modelli HYDRIM C61wd G4 dotati di un sistema di pulizia dei lumi. Se l’apparecchio HYDRIM è abilitato per SPL, sulla parete posteriore interna di sinistra della camera è presente un tappo bianco. Rack per strumenti verticali (optional) Spinotto SPL (incluso) Adattatore SPL e rack (inclusi) Montaggio dell’adattatore SPL 1 Retrarre l’anello di adattamento a innesto rapido, inserire l’attacco della linea pneumatica e rilasciare l’anello in modo che si blocchi. 2 3 Togliere la vite ed estrarre il tappo bianco presente sulla parete posteriore della camera dell’unità. 4 Collegare il tappo SPL al carrello HYDRIM. (è in posizione abbassata quando il rack SPL non viene utilizzato) Collegare l’adattatore SPL e il rack al carrello, agganciando la linguetta presente nella parte superiore dello spinotto SPL. Posizionare il cestello di pulizia cannule sui binari del rack SPL. Collegamento dei manipoli all’adattatore SPL 1 2 3 Inserire l’attacco per manipoli nel manipolo. Infilare il manipolo in una delle prese. Premere la linguetta per lo sgancio Multiflex Avvitare il raccordo LCS al corrispondente attacco per manipoli. Intramatic Utilizzo del menu touchscreen L’apparecchio HYDRIM rileverà automaticamente la presenza dell’adattatore SPL. Selezionare il programma desiderato e seguire le istruzioni. Al termine della procedura, aprire il portello e rimuovere i manipoli dall’adattatore spingendo la linguetta a lato del portello. Esempio: ciclo regolare 1 2 3 Notare che: come raccomandato dai produttori/fornitori leader di manipoli, le parti cave degli strumenti devono essere asciugate immediatamente dopo la pulizia e la disinfezione con aria compressa per uso medico, prima di effettuare ulteriori processi o la sterilizzazione. Quindi, effettuare le regolari procedure di manutenzione raccomandate e applicare gli agenti/oli di manutenzione. Si prega di seguire tutti i regolamenti locali in vigore in materia di salute e sicurezza. Italian Manutenzione e risoluzione dei problemi PULIZIA DEL FILTRO ACQUA SPL 1 2 3 4 Svitare il tappo sul lato anteriore del cilindro adattatore SPL. Togliere il filtro. Lavare il filtro e, se necessario, pulirlo con una spazzola morbida o con aria compressa. Controllare e pulire l’O-ring. Togliere eventuali impurità presenti all’interno dell’alloggiamento. Reinserire il filtro nell’alloggiamento e chiudere saldamente il tappo. SOSTITUZIONE DEL FILTRO BATTERIOLOGICO Sostituire il filtro batteriologico ogni 500 cicli o se è sporco. Per sostituire il filtro: 1. Spegnere l’unità, staccare il tubo A e togliere il filtro dalla staffa. 2. Scollegare il tubo B dalla parte posteriore del filtro. Tubo A Tubo B (dietro) 3. Prima di installare il filtro di ricambio (01-102119S) verificare che il senso della freccia sul filtro corrisponda al senso della freccia sulla staffa. (La freccia dovrebbe indicare a destra). 4. Inserire il filtro nella staffa e collegare il tubo B. Ricollegare il tubo A. 5. Dopo aver sostituito il filtro, entrare nella schermata “Resettare contat. SPL” nel menu Setup e resettare il contatore a 0. Messaggi sistema LCS Significato Rimedio – Utente Sostituire filtro acqua Il filtro acqua nell’adattatore SPL è sporco o intasato. • Seguire le istruzioni precedentemente menzionate: “Pulizia del filtro acqua .” Sostituire filtro aria Il filtro aria batteriologico nel fermapiede è sporco. • Seguire le istruzioni precedentemente menzionate: “Sostituzione del filtro dell’aria.” • Controllare se l’adattatore SPL è correttamente posizionato. • È stato selezionato l’SPL all’avvio del programma, ma l’SPL non è installato. CF 12 Filtro acqua intasato Problema con il sistema di pulizia dei lumi. • Sostituire il filtro acqua, vedere “Pulizia del filtro acqua.” • Se la pulizia non risolve il problema, sostituire il filtro acqua (01-113742S) • Se il messaggio persiste, rivolgersi all’assistenza. CF 18 Sensore pressione filtro acqua bloccato Problema con il sistema di pulizia dei lumi. • Rivolgersi all’assistenza. CF 26 Errore valvola aria Problema con la pressione di alimentazione dell’aria nella linea pneumatica. • Rivolgersi all’assistenza. CF 30 Adattatore SPL scollegato Incoming pressione di alimentazione di aria deve essere compresa tra 1 bar (15 psi)-10 bar (145 psi) • Controllare che l’adattatore SPL sia correttamente posizionato. Problema con il sistema di pulizia dei lumi. Chinese HYDRIM ® C61wd G4 内腔清洗系统 (LCS) 内腔清洗系统 (LCS) 接头只能用 于配备内腔清洗系统的 HYDRIM C61wd G4 机型。 如果您的 HYDRIM能够配合LCS使用,您 将会在腔内左后方上看到白色插 头。 垂向器械网架(可选) LCS 插头 LCS 插头和网架 (包括) (包括) 设置 LCS 插头 1 回拉快速接头连接环,插 入空气管道接头并释放 环,以便锁定。 2 3 4 将 LCS 插头连接到 HYDRIM 滚轮。 (未使用 LCS 网架时保持向下位置) 在设备腔后壁上拆下 螺丝并拔出白色插头 将 LCS 插头和网架连接到滚轮,在 LCS 插头的顶部占用选项。 将套管 清洗篮置于 LCS 网架的栏杆上。 将手机与 LCS 插头连接 1 2 3 将手机接头插入手机。 将手机推入一个端口中。 按下选项以 释放 Multiflex 拧紧 LCS,安装到合适 的手机接头。 Intramatic 使用触摸屏菜单 举例: 常规循环 HYDRIM 将会自动检测 LCS 插头的存在。 选择 所需的程序,并按照提示 操作。 完成后,打开设 备门,通过在端口侧面推 按选项取出手机。 1 2 3 请注意:根据领先的机头制造 商/供应商的建议,仪器中空 部分在清洗和消毒之后应该立 即使用医用压缩空气进行干 燥,然后才能进行额外的处理 或灭菌。在这之后,应该执行 推荐的例行维护程序,并施加 保养剂/保养油。请遵循您所 在地区所有适用的卫生与安全 监管规定。 Chinese 保养与故障排除 清洁 LCS 水过滤器 1 2 3 4 在 LCS 插头圆筒的前方 旋开盖子。 取下过滤器。 冲洗过滤器,如果需要,用软刷或用压 缩空气进行清洁。 检查并清洁 O 型密 封圈。 清除壳体内部的任何残余物。 将过滤器重新装到壳体 上并紧紧关闭盖子。 更换空气过滤器 每 500 个循环或如果已脏, 则更换空气过滤器。 要更换过滤器: 1. 关闭设备电源并断开 A 管, 然后从托架上取出过滤器。 2. 从过滤器的后面断开 B 管。 A管 3. 在安装替代的过滤器 (01-102119S) 之前, 请检查过滤器机器上的箭头标记与托架上 的箭头方向相符。 (箭头应当指向右侧)。 4. 将过滤器插入托架并连接 B 管。重新连接 A 管。 5. 在更换过滤器后,转到设置菜单中的“重置 LCS 计数器”屏幕并且将计数器重置为 0。 B 管(后面) LCS 系统消息 这意味着什么 更换水过滤器 LCS 中的水过滤器已脏或者堵塞。 • 遵照上面的说明:“清洁 LCS 水过滤器。” 更换空气过滤器 踢板中的空气过滤器已脏。 • 遵照上面的说明: “更换细菌空气过滤器。” 怎么办 - 用户 • 检查 LCS 插头是否正确定位。 •  您在程序启动处选择了 LCS 但是 LCS 未安装。 CF 12 水 内腔清洗系统问题。 • 更换水过滤器,参见 “清洁 LCS 水过滤器。” • 如果清洁并未解决问题,请更换水过滤器 过滤器阻塞 (01-113742S) • 如信息仍存在,请致电服务部。 CF 18 水过滤器 内腔清洗系统问题。 • 请致电经销商。 空气管道中供气压力问题。 • 请致电经销商。 压力传感器卡住 CF 26 空气阀错误 输入的空气供应压力必须在 1 bar (15 psi)至 10 bar (145 psi)之间。 CF 30 LCS 插头 断开 内腔清洗系统问题。 • 检查 LCS 插头是否正确定位。 SciCan Ltd. 1440 Don Mills Road Toronto, Ontario M3B 3P9 Tel: 416-445-1600 Toll free: 1-800-667-7733 0123 Fax: 416-445-2727 EU Rep: SciCan GmbH Wangener Strasse 78 88299 Leutkirch, GERMANY Tel.: +49 (0)7561 98343 - 0 Fax: +49 (0)7561 98343 - 699 SciCan Medtech AG Alpenstrasse 16 6300 Zug, SWITZERLAND Tel.: +41 (0) 41-727 7027 Fax.: +41 (0) 41-727 7029 www.scican.com SciCan SciCanTV @SciCan HYDRIM is a registered trademark and Your Infection Control Specialist is a trademark of SciCan Ltd. 95-113906 EU EFIGSC R2 Manufactured by
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14

SciCan Hydrim C61wd G4 Manual de usuario

Tipo
Manual de usuario