Ohaus FC5306 Manual de usuario

Tipo
Manual de usuario

Este manual también es adecuado para

Frontier
TM
Centrifuge
FC5306
Instruction Manual
Centrifuga Frontier
TM
FC5306
Manual de instrucciones
Centrifugeuse Frontier
TM
FC5306
Manuel d’instructions
Frontier
TM
-Zentrifuge
FC5306
Bedienungsanleitung
Centrifuga Frontier
TM
FC5306
Manuale di istruzioni
TABLE OF CONTENTS
1 INTRODUCTION .............................................................. 1
1.1 Safety measures .................................................................................. 1
2 INSTALLATION................................................................ 3
2.1 Package contents ................................................................................. 3
2.2 Replacement or Installation of Rotor .................................................... 4
2.3 Power On ............................................................................................. 5
2.4 Preparation for Centrifugation .............................................................. 5
2.5 Centrifugation (Running) ...................................................................... 5
3 SPECIAL ATTENTION ..................................................... 6
3.1 Power supply ........................................................................................ 6
3.2 Use and maintenance .......................................................................... 6
4 SPECIFICATIONS ............................................................ 7
5 COMPLIANCE .................................................................. 8
Frontier
TM
FC5306 EN-1
1 INTRODUCTION
Thank you for choosing this OHAUS product.
Please read the manual completely before using the Frontier
TM
FC5306 to
avoid incorrect operation.
You will find much helpful functionality in Frontier
TM
5000 series centrifuge.
A few of the many existing features for Frontier
TM
FC5306 are:
Small and exquisite body, one-key operation
Convenient and quick installation and replacement of rotor without
need of using tool
Rotor will immediately stop running if lid is opened during operation
Desktop power adapter is applicable to 100-240V AC power supply
Elaborate product with low noise and long service time
1.1 Safety measures
CAUTION: Read all safety warnings before installing, making
connections, or servicing this equipment. Failure to comply with
these warnings could result in personal injury and/or property
damage. Retain all instructions for future reference.
Measures for your protection
Never work in an environment subject to explosion hazards!
The housing of the instrument is not gas tight. (explosion
hazard due to spark formation, corrosion caused by the
ingress of gases)
When using chemicals and solvents, comply with the
instructions of the producer and the general lab safety rules.
Measures for operational safety
Do not lean on the centrifuge.
Do not stay within the clearance envelope longer than necessary for
operational reasons.
Do not deposit any potentially hazardous materials within the clearance
envelope.
Do not unscrew the two halves of the housing.
Dry off any liquid spills immediately! The instrument is not watertight.
EN-2 Frontier
TM
FC5306
Verify that the AC adapter’s input voltage range and plug type are
compatible with the local AC power to be used.
Do not position the equipment such that it is difficult to disconnect the
AC adapter from the power receptacle.
Make sure that the power cord does not pose a potential obstacle or
tripping hazard.
The equipment is for indoor use only. Use the equipment only in dry
locations.
Use only approved accessories and peripherals.
Operate the equipment only under ambient conditions specified in
these instructions.
Disconnect the equipment from the power supply when cleaning.
Do not operate the equipment in hazardous or unstable environments.
Service should only be performed by authorized personnel.
Exclude the following environmental influences:
Powerful vibrations
Direct sunlight
Atmospheric humidity greater than 80%
Corrosive gases present
Temperatures below 5 °C and above 40 °C
Powerful electric or magnetic fields
Frontier
TM
FC5306 EN-3
2 INSTALLATION
Carefully unpack the box.
2.1 Package contents
The model Frontier
TM
FC5306 has the following items:
Frontier
TM
FC5306
Photo
Units
Frontier
TM
FC5306 main
machine
1 set
Angle rotor ( 8 x 1.5/2ml)
1 set
4 x 0.2ml rectangle rotor
for PCR 8-tube bank
1 set
Dedicated DC24V power
adapter
1 set
Eight 0.5 ml adapters and
eight 0.2 ml adapters
8 x 2 units
Manual
1 set
EN-4 Frontier
TM
FC5306
2.2 Replacement or Installation of Rotor
1. When the machine is delivered from the factory, the seizing shaft
sleeve and angle rotor are firmly fixed on the shaft of the drive motor.
2. If it is necessary to dismantle an installed rotor, the user should hold
the outside edge of the rotor, and push the protruding clamp towards
its center with two thumbs, then lift the rotor upwards.
3. To reinstall rotor, align the oval hole of rotor over the clamp, then
gently press down the rotor until the clamp clicks into place.
角转头
电机轴套
Angle Rotor
Motor Shaft Sleeve
Frontier
TM
FC5306 EN-5
2.3 Power On
1. With the power switch on the back of this machine OFF.
2. Connect the output end of the adapter to the power socket on the
back of the machine, and then plug the unit to a power source.
2.4 Preparation for Centrifugation
1. Open the transparent lid (see photo below) and properly add load
in a balanced manner (refer to section “3.2 use and maintenance). (refer to “section 3.2 use and maintenance”).
2. Close the lid then press the power switch on the back of this
machine to ON (note: The rotor will not run as long as the lid is open).
2.5 Centrifugation (Running)
1 After the lid is closed and the machine switched on, the rotor will
automatically start running with its speed increasing to the rated
speed, and then it will run stably.
2. Open the lid when the requirement of centrifugation is met, the rotor
will immediately slow down to a complete stop.
3. It is unnecessary to turn off power if another centrifugation is
required.
4. Turn off and unplug the machine when all centrifugation work is
completed or when not in use.
press down to
open the lid
EN-6 Frontier
TM
FC5306
3 SPECIAL ATTENTION
3.1 Power supply
1. The power supply to this machine must be realized through the
adapter provided together with the machine. The AC plug of adapter
should not be connected with three-phase power source.
2. Don’t insert or disconnect the power plug of the adapter with wet hand.
Don’t forcibly pull the cable to disconnect the power plug.
3.2 Use and maintenance
1. Prepare the micro burette of standard specification with hinged
cover for centrifugation. It is only allowed to add sample till approx.
70% of nominal capacity at most. Visually check balance during
adding of sample. Don’t detach any poisonous or infectious biological
sample. Dont detach flammable or explosive chemical sample.
2. The centrifugal tubes with sample must be symmetrically placed into
the centrifugal tube holes of the rotor to keep the rotor balanced.
3. When the rotor is installed or replaced, excessive sidewise force
should be avoided to prevent its deformation and bending. If the
seizing shaft sleeve around the motor shaft is a bit loose, it is
necessary to fasten the M3 set screw again.
4. Do not use broken or deformed rotor.
5. During daily cleaning and maintenance, the machine should be wiped
with clean water or neutral detergent, and be dried through natural
ventilation.
6. During use, attention should be paid to symmetrical and horizontal
placement of centrifugal tubes.
WARNING: Disconnect the AC adapter from the AC power
supply and the equipment before cleaning. Do not allow liquids
to enter the AC adapter or the equipment.
Frontier
TM
FC5306 EN-7
4 SPECIFICATIONS
The technical data is valid under the following ambient conditions:
Operating temperature: 5°C to 40°C
Relative humidity: 20% to 85% relative humidity, non-condensing
Height above sea level: Up to 2000 m
Pollution degree: 2
Installation category: II
Main supply voltage fluctuations: up to ± 10% of the nominal voltage
Model
Frontier Mini CentrifugeFC5306
Rotor
Angle Rotor
Tube Bank Rotor
Rotating Speed Scope
6000 rpm ± 5%
Centrifugal Force
2000 g ± 10%
Rotating Direction
Counter clockwise
Max Capacity
8 x 1.5 ml / 2 ml
4 x 0.2 ml-PCR 8-tube
bank
or 32 x 0.2 ml single
centrifugal tube
Adapter
0.2 ml / 0.5 ml
-
Accelerating Time
≤ 2 s
Decelerating Time
≤ 3 s
Drive Mode and Motor
Direct drive, with low noise attributable to rolling bearing
Power Supply
Input: 100-240V ~ 50/60Hz 1.5A
Output: 24V .6A
Dimensions (W × D × H)
5.9 × 6.6 × 5.0 in. / 150 × 168 × 128 mm
Net Weight
2.316 lb / 1.05 Kg (including free-load angle rotor)
Shipping Dimensions
(W × D × H)
8.5 × 12.0 × 7.6 in. / 216 × 305 × 194 mm
Shipping Weight
4.079 lb / 1.85 Kg
EN-8 Frontier
TM
FC5306
5 COMPLIANCE
Compliance to the following standards is indicated by the
corresponding mark on the product.
Marking
Standard
This product conforms to the EMC directive 2004/108/EC and
the Low Voltage Directive 2006/95/EC. The complete
Declaration of Conformity is available online at
http://europe.ohaus.com/europe/en/home/support/compliance/c
e-declaration-of-conformity.aspx
Disposal
In conformance with the European Directive 2002/96/EC on
Waste Electrical and Electronic Equipment (WEEE) this
device may not be disposed of in domestic waste. This also
applies to countries outside the EU, per their specific
requirements.
Please dispose of this product in accordance with local
regulations at the collecting point specified for electrical and
electronic equipment.
If you have any questions, please contact the responsible
authority or the distributor from which you purchased this
device.
Should this device be passed on to other parties (for private
or professional use), the content of this regulation must also
be related.
For disposal instructions in Europe, refer to
http://europe.ohaus.com/europe/en/home/support/weee.aspx.
Thank you for your contribution to environmental protection.
Frontier
TM
FC5306 EN-9
FCC Note
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference
in a residential installation. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Industry Canada Note
This Class B digital apparatus complies with Canadian ICES-001.
TABLA DE CONTENIDO
1 INTRODUCCIÓN .............................................................. 1
1.1 Medidas de seguridad .......................................................................... 1
2 INSTALACIÓN ................................................................. 3
2.1 Contenido del paquete ......................................................................... 3
2.2 Reemplazo o instalación del rotor ........................................................ 4
2.3 Encendido ............................................................................................ 5
2.4 Preparación para la centrifugación ...................................................... 5
2.5 Centrifugación (operación) ................................................................... 5
3 ATENCIÓN ESPECIAL .................................................... 6
3.1 Alimentación eléctrica .......................................................................... 6
3.2 Uso y mantenimiento ........................................................................... 6
4 ESPECIFICACIONES ....................................................... 7
5 CONFORMIDAD NORMATIVA ........................................ 8
Frontier
TM
FC5306 ES-1
1 INTRODUCCIÓN
Gracias por elegir este producto OHAUS.
Lea este manual completamente antes de usar la Frontier
TM
FC5306 para
evitar la operación incorrecta.
Encontrará mucha funcionalidad útil en la centrífuga Frontier
TM
serie 5000.
Algunas de las muchas características actuales de la Frontier
TM
FC5306
son:
Cuerpo pequeño y refinado, y operación con una tecla
Instalación conveniente y rápida y reemplazo del rotor sin necesidad
de usar herramientas
El rotor se detiene inmediatamente si se abre la tapa durante la
operación
El adaptador de corriente de mesa es compatible con la corriente
eléctrica de 100 a 240 V
Bajo nivel de ruido durante la operación.
1.1 Medidas de seguridad
PRECAUCIÓN: Lea todas las advertencias de seguridad antes
de instalar, hacer conexiones o dar servicio a este equipo. La
falta de cumplimiento de estas advertencias puede resultar en
lesiones personales o daños materiales. Conserve todas las
instrucciones para referencia futura.
Medidas para su protección
Nunca trabaje en un ambiente sujeto a peligros de
explosión El equipo no es impermeable al gas.
(Peligro de explosión debido a formación de chispas;
corrosión debido al ingreso de gases).
Cuando use químicos y solventes, siga las instrucciones del
fabricante y reglas de seguridad generales de laboratorio.
ES-2 Frontier
TM
FC5306
Medidas para seguridad operativa
No se apoye sobre la centrífuga.
No permanezca dentro del espacio libre de seguridad más tiempo del
necesario por razones operativas.
No deposite ninguna sustancia peligrosa dentro del espacio libre de
seguridad.
No destornille la tapa del equipo.
Seque de inmediato cualquier líquido que se derrame. El instrumento
no es hermético al agua.
Verifique que el rango del voltaje de entrada del adaptador de corriente
y el tipo de enchufe sean compatibles con la corriente alterna local que
va a usarse.
No coloque el equipo de tal forma que sea difícil desconectar el
adaptador de corriente de la toma de corriente.
Verifique que el cordón eléctrico no represente un obstáculo potencial
o riesgo de tropezarse.
Este equipo es solamente para uso en interiores. Use el equipo
solamente en lugares secos.
Use solo accesorios y periféricos aprobados.
Opere el equipo solo en las condiciones ambientales especificadas en
estas instrucciones.
Desconecte el equipo de la alimentación eléctrica antes de limpiarlo.
No opere el equipo en ambientes peligrosos o inestables.
El servicio debe proporcionarse solamente por personal autorizado.
Excluya las siguientes influencias del medio ambiente:
Vibraciones potentes
Luz solar directa
Humedad atmosférica mayor del 80%
Presencia de gases corrosivos
Temperaturas menores de 5 °C y mayores de 40 °C
Campos eléctricos o magnéticos potentes
Frontier
TM
FC5306 ES-3
2 INSTALACIÓN
Desempaque con cuidado la caja.
2.1 Contenido del paquete
El modelo Frontier
TM
FC5306 tiene los siguientes componentes:
Frontier
TM
FC5306
Imagen
Unidades
Frontier
TM
FC5306
Equipo principal
1
Rotor angular
(8 x 1.5/2 ml)
1
Rotor rectangular 4 x 0.2
ml para banco PCR de
8 tubos
1
Adaptador de corriente
de 24 VCD
1
Ocho adaptadores de
0.5 ml y ocho
adaptadores de 0.2 ml
8 x 2 unidades
Manual
1
ES-4 Frontier
TM
FC5306
2.2 Reemplazo o instalación del rotor
1. Cuando el equipo se entrega de fábrica, el sujetador y el rotor angular
se fijan firmemente en el eje del motor.
2. Si es necesario reemplazar un rotor instalado, el usuario deberá
sostener el borde externo del rotor y empujar la abrazadera
sobresaliente hacia el centro con los dos pulgares, y luego mover el
rotor hacia arriba.
3. Para instalar el rotor, alinie el orificio ovalado del rotor sobre la
abrazadera y luego presione hacia abajo el rotor hasta que la
abrazadera haga clic en su lugar.
角转头
电机轴套
Rotor angular
Sujetador
Abrazadera
Frontier
TM
FC5306 ES-5
2.3 Encendido
1. El interruptor de corriente en la parte posterior del equipo debe estar
en OFF (Apagado).
2. Conecte el extremo de salida del adaptador en el enchufe de
corriente de la parte posterior del equipo, y luego conecte la unidad
en una fuente de energía.
2.4 Preparación para la centrifugación
1. Abra la tapa transparente (vea la foto siguiente) y agregue carga
apropiadamente en una forma balanceada (consulte la sección “3.2
Uso y mantenimiento”). (Consulte la “sección 3.2 Uso y mantenimiento”).
2. Cierre la tapa y luego presione el interruptor de corriente en la parte
posterior de la máquina a la posición ON (Encendido) (Nota: El rotor
no funcionará mientras la tapa esté abierta).
2.5 Centrifugación (operación)
1 Después de cerrar la tapa y de encender el equipo, el rotor
comenzará a funcionar automáticamente y la velocidad se
incrementará hasta la velocidad nominal, y luego funcionará de
manera estable.
2. Abra la tapa cuando se cumpla el requerimiento de centrifugación, y
el rotor reducirá su velocidad inmediatamente hasta detenerse por
completo.
3. Para realizar otra centrofugación, solo cierre la tapa nuevamente.
4. Apague y desenchufe la máquina cuando haya terminado todo el
trabajo de centrifugación o cuando no esté en uso
Presione para
abrir la tapa
ES-6 Frontier
TM
FC5306
3 ATENCN ESPECIAL
3.1 Alimentacn eléctrica
1. La alimentación eléctrica hacia este equipo debe hacerse a través del
adaptador que se proporciona con el equipo. El enchufe de corriente
del adaptador no deberá conectarse con la alimentación eléctrica
trifásica.
2. No inserte ni desconecte el enchufe de corriente del adaptador con las
manos húmedas.
No jale el cable con fuerza para desconectar el enchufe de corriente.
3.2 Uso y mantenimiento
1. Prepare la microbureta de especificación estándar con cubierta de
bisagras para centrifugación. Solo se permite agregar muestra Un
70% de la capacidad nominal. No ingrese ninguna muestra biológica
venenosa o infecciosa. No ingrese ninguna muestra química
inflamable o explosiva.
2. Los tubos centrífugos con muestra deben colocarse en forma
simétrica en los orificios del rotor para mantener el
rotor balanceado.
3. Al instalar o reemplazar el rotor, se deberá evitar ejercer fuerza
lateral excesiva para prevenir la deformación y flexión. Si el
sujetador de agarre alrededor del eje del motor está un
poco flojo, es necesario apretar nuevamente el tornillo de sujeción M3.
4. No use el rotor si está roto o deformado.
5. Durante la limpieza y mantenimiento diario, el equipo debe frotarse
con agua o detergente neutro, y debe secarse mediante
ventilación natural.
6. Durante el uso, se debe prestar atención a la colocación simétrica
y horizontal de los tubos centrífugos.
ADVERTENCIA: Desconecte el adaptador de corriente de la
alimentación eléctrica y del equipo antes de la limpieza. No
permita que entren líquidos al adaptador de corriente ni al
equipo.
Frontier
TM
FC5306 ES-7
4 ESPECIFICACIONES
Los datos técnicos son válidos en las siguientes condiciones ambientales:
Temperatura de operación: 5 °C a 40 °C
Humedad relativa: humedad relativa de 20% a 85%, no condensante
Altura sobre el nivel del mar: hasta 2000 m
Grado de contaminación: 2
Categoría de instalación: II
Fluctuaciones de voltaje de la alimentación principal: hasta ± 10% del
voltaje nominal
Modelo
Minicentrífuga FrontierFC5306
Rotor
Rotor angular
Rotor de tubos
Alcance de la
velocidad de rotación
6000 rpm ± 5%
Fuerza centrífuga
2000 g ± 10%
Dirección de rotación
En sentido contrario a las manecillas del reloj
xima capacidad
8 x 1.5 ml / 2 ml
4 x 0.2 ml - Banco de 8 tubos PCR
o tubo centrífugo simple de 32 x 0.2
ml
Adaptador
0.2 ml / 0.5 ml
-
Tiempo de aceleración
≤ 2 s
Tiempo de
desaceleración
≤ 3 s
Accionamiento del
motor
Accionamiento directo
Alimentación eléctrica
Entrada: 100-240 V ~ 50/60Hz 1.5 A
Salida: 24 V .6A
Dimensiones
150 mm An x 168 mm Pr x 128 mm Al
5.9 pulg. An x 6.6 pulg. Pr x 5.0 pulg. Al
Peso neto
1.05 kg (incluyendo rotor angular)
2.316 lb
Dimensiones del envío
8.5 × 12.0 × 7.6 in. / 216 × 305 × 194 mm
Peso de envío
4.079 lb / 1.85 Kg
ES-8 Frontier
TM
FC5306
5 CONFORMIDAD NORMATIVA
La marca correspondiente en el producto indica la conformidad con
los siguientes estándares.
Marca
Estándar
Este producto está en conformidad con la directiva EMC
2004/108/EC y con la directiva para Voltaje Bajo 2006/95/EC.
La Declaración de conformidad completa está disponible en
línea en
http://europe.ohaus.com/europe/en/home/support/compliance/c
e-declaration-of-conformity.aspx
Disposición final
De conformidad con la Directiva Europea 2002/96/EC sobre
Residuos de Equipos Eléctricos y Electrónicos (WEEE), este
dispositivo no puede desecharse con la basura doméstica.
Esto también es aplicable para países fuera de la UE, según
sus requerimientos específicos.
Deseche este producto de acuerdo con las regulaciones
locales en el punto de recolección especificado para equipos
eléctricos y electrónicos.
Si tiene alguna pregunta, comuníquese con la autoridad
responsable o con el distribuidor a quien compró este
dispositivo.
En caso de que este dispositivo sea transferido a otras
partes (para uso privado o profesional), también deberá
mencionarse el contenido de esta regulación.
Para consultar las instrucciones de eliminación en Europa,
visite
http://europe.ohaus.com/europe/en/home/support/weee.aspx.
Gracias por su contribución a la protección ambiental.
Frontier
TM
FC5306 ES-9
Nota de la FCC
Este equipo ha sido probado y se encontró que cumple con los límites para
un dispositivo digital clase B, consecuente con la Parte 15 de las Reglas
de la FCC. Estos límites están diseñados para proporcionar una protección
razonable contra interferencia dañina cuando el equipo es operado en un
ambiente comercial. Este equipo genera, usa y puede irradiar frecuencias
de radio y, si no es instalado y utilizado de acuerdo con el manual de
instrucciones, puede causar interferencia dañina a las
radiocomunicaciones. Sin embargo, no hay garantía de que no se
produzcan interferencias en una instalación en particular. Si este equipo
causa interferencias perjudiciales en la radio o televisión, lo cual puede
comprobarse encendiéndolo y apagándolo, se recomienda al usuario que
intente corregir la interferencia mediante una o más de las siguientes
medidas:
Reorientar o reubicar la antena receptora.
Aumentar la separación entre el equipo y el receptor.
Conectar el equipo a un tomacorriente en un circuito diferente de aquel al
que el receptor está conectado.
Consulte al distribuidor o a un técnico de radio / TV con experiencia para
obtener ayuda.
Nota de Industry Canada
Este aparato digital Clase B cumple con la norma canadiense ICES-001.
TABLE DES MATRES
1 INTRODUCTION .............................................................. 1
1.1 Mesures de sécurité ............................................................................. 1
2 INSTALLATION................................................................ 3
2.1 Contenu de l’emballage ....................................................................... 3
2.2 Remplacement ou installation du rotor ................................................ 4
2.3 Mise sous tension ................................................................................ 5
2.4 Préparation à la centrifugation ............................................................. 5
2.5 Centrifugation (en fonctionnement) ...................................................... 5
3 PRÉCAUTIONS PARTICULIÈRES .................................. 6
3.1 Alimentation ......................................................................................... 6
3.2 Utilisation et maintenance .................................................................... 6
4 SPÉCIFICATIONS ............................................................ 7
5 CONFORMITÉ .................................................................. 8
Frontier
TM
FC5306 FR-1
1 INTRODUCTION
Merci d'avoir choisi ce produit OHAUS.
Veuillez lire le manuel intégralement avant d'utiliser la centrifugeuse
Frontier
TM
FC5306 afin d'éviter un fonctionnement incorrect.
Un grand nombre de fonctions utiles de la centrifugeuse de la série
Frontier
TM
5000 sont à découvrir. Veuillez trouver quelques-unes parmi les
nombreuses caractéristiques actuelles de la série Frontier
TM
5000:
Petit Boitier, beau désigne, faible encombrement et une seule
touche de fonctionnement
L'installation (et /ou) le remplacement de rotor est facile , rapide,
sans avoir besoin d'aucun outil
Le rotor s'arrêtera immédiatement si le couvercle est soulevé
pendant le fonctionnement
L'adaptateur d'alimentation fonctionne dans la plage de 100 à 240 V
CA
Produit élaboré au bruit limité et à longue durée d'utilisation
1.1 Mesures de sécurité
PRÉCAUTIONS : Lire tous les avertissements de sécurité avant
de procéder à l’installation, aux connexions ou à la maintenance
de cet équipement. Si ces avertissements ne sont pas
respectés, des accidents et/ou des dommages matériels
peuvent en résulter. Conserver toutes les instructions pour référence
ultérieure.
Mesures de protection à respecter
Ne jamais travailler dans un environnement pouvant
présenter des dangers d'explosion ! Le boîtier de
l'instrument n'est pas étanche aux gaz (danger d'explosion
due à la formation d'étincelles, corrosion provoquée par
l'introduction de gaz).
Avec des produits chimiques et des solvants, se conformer
aux instructions du fabricant des produits chimiques et aux
règles générales de sécurité du laboratoire.
FR-2 Frontier
TM
FC5306
Mesures à respecter pour la sécurité du fonctionnement
Ne pas s'appuyer contre la centrifugeuse.
Ne pas rester dans la zone de sécurité autour de la centrifugeuse plus
que nécessaire pour des raisons d'exploitation.
Ne laisser aucun produit potentiellement dangereux dans la zone de
sécurité autour de la centrifugeuse.
Ne pas dévisser les deux moitiés de l'enceinte.
Sécher immédiatement tout déversement de liquide. L'instrument n'est
pas étanche.
Vérifier que la plage de tension d'entrée de l'adaptateur CA et le type
de prise sont compatibles avec l'alimentation alternative locale devant
être utilisée.
Ne pas positionner l'équipement pour qu'il soit difficile de déconnecter
l'adaptateur CA de la prise d'alimentation.
S'assurer que le cordon d'alimentation n’est pas un obstacle potentiel
ou ne présente aucun danger de chute.
L'équipement ne doit être utilisé qu'en intérieur. Utiliser l'équipement
seulement dans des lieux secs.
Utiliser uniquement les accessoires et les périphériques approuvés.
Faire fonctionner l'équipement uniquement dans les conditions
ambiantes spécifiées dans ces instructions.
Déconnecter l'équipement de l'alimentation électrique pour le nettoyer.
Ne pas utiliser l'équipement dans des environnements dangereux ou
instables.
La maintenance doit être exclusivement exécutée par un personnel
autorisé.
Exclure les influences environnementales suivantes :
Vibrations puissantes
Exposition au Rayon de soleil
Humidité atmosphérique supérieure à 80 %
Présence de gaz corrosifs
Températures supérieures à 40 °C et inférieures à 5 °C
Champs électriques ou magnétiques puissants
Frontier
TM
FC5306 FR-3
2 INSTALLATION
Déballer soigneusement la boîte.
2.1 Contenu de l’emballage
Le modèle Frontier
TM
FC5306 contient les articles suivants :
Frontier
TM
FC5306
Photo
Unités
Machine principale
Frontier
TM
FC5306
1 ensemble
Rotor angulaire ( 8 x
1.5/2ml)
1 ensemble
Rotor rectangulaire de 4 x
0.2 ml pour un ensemble
de tubes PCR 8
1 ensemble
Adaptateur dédié
d'alimentation continue
24 V
1 ensemble
Huit adaptateurs 0.5 ml et
huit adaptateurs 0.2 ml
8 x 2 unités
Manuel
1 ensemble
FR-4 Frontier
TM
FC5306
2.2 Remplacement ou installation du rotor
1. Lorsque la machine est livrée depuis l'usine, le manchon de l'axe de
saisie et le rotor angulaire sont fermement fixés sur l'axe du moteur
d'entraînemen.
2. S'il s'avère nécessaire de démonter un rotor installé, l'utilisateur doit
maintenir le bord extérieur du rotor et pousser sur la protubérance du
collier de serrage vers son centre avec les deux pouces, et ensuite
soulever le rotor.
3. Pour réinstaller le rotor, aligner le trou ovale de rotor par-dessus le
collier de serrage puis Appuyez doucement sur le rotor jusqu'à ce
que le collier s'enclenche.
角转头
电机轴套
Rotor angulaire
Manchon de l'arbre
moteur
Collier de
serrage
Frontier
TM
FC5306 FR-5
2.3 Mise sous tension
1. Connecter la sortie de l'adaptateur sur la prise à l'arrière de la
machine et brancher l'unité à une source d'alimentation.
2. Mettre sous tension avec le bouton MARCHE ARRET à l'arrière de
la centrifugeuse.
2.4 Préparation à la centrifugation
1. Ouvrir le couvercle transparent (voir la photo ci-dessous) et
introduire la charge correctement de manière équilibrée (reportez-
vous à la section 3.2, Utilisation et maintenance).
2. Refermer le couvercle et appuyer sur le contacteur d'alimentation
MARCHE à l'arrière de cette machine (remarque : le rotor ne
fonctionnera pas tant que le couvercle est ouvert).
2.5 Centrifugation (en fonctionnement)
1 Le couvercle étant fermé et la machine sous tension, le rotor
commencera automatiquement sa rotation à vitesse croissante
jusqu'à sa valeur nominale, puis se stabilisera.
2. Ne pas ouvrir le couvercle de la centrifugeuse tant que la
centrifugation n'est pas finie et attendre jusqu'à l'arrêt complet du
rotor.
3. Il n'est pas nécessaire de mettre l'appareil hors tension si une autre
centrifugation s'avère nécessaire.
4. Mettre hors tension et débrancher la machine lorsque tout le travail
de centrifugation est terminé ou lorsqu'elle n'est pas en utilisation.
Appuyer sur le
couvercle
pour l'ouvrir
FR-6 Frontier
TM
FC5306
3 PRÉCAUTIONS PARTICULIÈRES
3.1 Alimentation
1. L'alimentation de cette machine doit s'effectuer avec l'adaptateur
l'accompagnant. La prise CA de l'adaptateur ne doit pas être
connectée sur une alimentation triphasée.
2. Ne Brancher, ni Débrancher la prise d'alimentation de
l'adaptateur avec une main mouillée. Ne pas tirer sur le câble pour le
débrancher de la prise.
3.2 Utilisation et maintenance
1. L'ajout d'un échantillon n'est autorisé que jusqu'à environ 70 % de la
capacité nominale.Vérifier visuellement l'équilibrage lors de l'ajout d'un
échantillon. Ne pas segmenter aucun échantillon biologique infectieux
ou toxique. Ne pas segmenter aucun échantillon de produits chimiques
explosifs ou inflammables.
2. Les tubes à essai avec un échantillon à centrifuger doivent être
positionnés symétriquement dans les orifices, afin de conserver
l'équilibrage du rotor
3. Lorsque le rotor est installé ou remplacé, toute force latérale excessive
doit être évitée afin d'empêcher sa déformation ou son cintrage. Si le
manchon de l'axe de saisie autour de l'axe du moteur d'entraînement
est légèrement desserré, il est nécessaire de resserrer la vis de
pression M3.
4. Ne jamais utiliser un rotor brisé ou déformé.
5. Au cours de la maintenance et du nettoyage quotidiens, la machine
doit être essuyée à l'eau propre et avec un détergent neutre et être
séchée par ventilation naturelle. .
6. Pendant l'utilisation, prendre toutes les précautions afin de conserver le
positionnement horizontal et symétrique des tubes de centrifugation.
AVERTISSEMENT : Déconnecter l'adaptateur CA de
l'alimentation secteur et débrancher l'équipement avant le
nettoyage. Ne laisser aucun liquide pénétrer dans l'adaptateur
CA ou dans l'équipement.
Frontier
TM
FC5306 FR-7
4 SPÉCIFICATIONS
Les données techniques sont valides dans les conditions ambiantes
suivantes :
Température de fonctionnement : 5 à 40 °C
Humidité relative : humidité relative entre 20 et 85 %, sans condensation
Altitude au-dessus du niveau de la mer : Jusqu'à 2000 m
Degré de pollution : 2
Catégorie d'installation : II
Variations secteur de ±10 % de la tension nominale
Modèle
Mini Centrifugeuse Frontier FC5306
Rotor
Rotor angulaire
Rotor réceptacle des tubes à essai
Spécifications de la
vitesse de rotation
6000 rpm ± 5%
Force centrifuge
2000 g ± 10%
Sens de la rotation
contre les aiguille d'une montre
Capacité maximum
8 x 1.5 ml / 2 ml
Rotor rectangulaire pour 8 PCR de
4 x 0.2 ml
32 x 0.2 ml tube de centrifugation à
usage unique
Adaptateur
0.2 ml / 0.5 ml
-
Durée de l'accération
≤ 2 s
Durée de la dération
≤ 3 s
Moteur et mode
d'entraînement
Entraînement direct, avec faible bruit, à cause des roulements
Alimentation
Entrée : 100-240 V ~ 50/60Hz 1.5 A
Sortie : 24 V
Dimensions
150 L x 168 P x 128 H mm
5.9 L x 6.6 P x 5 H po
Poids net
1.05 Kg (comprenant le rotor angulaire hors charge)
2.316 lbs.
Dimensions d'expédition
8.5 × 12.0 × 7.6 in. / 216 × 305 × 194 mm
Poids à l'expédition
4.079 lb / 1.85 kg
.6A
FR-8 Frontier
TM
FC5306
5 CONFORMITÉ
La conformité aux normes suivantes est indiquée par le marquage
correspondant sur le produit.
Marquage
Standard
Ce produit est conforme à la directive EMC 2004/108/EC et à la
directive de basse tension 2006/95/EC. L'ensemble de la
Déclaration de conformité est disponible en ligne sur
http://europe.ohaus.com/europe/en/home/support/compliance/c
e-declaration-of-conformity.aspx
Dechets
Conformément à la directive européenne 2002/96/CE sur les
déchets d’équipements électriques et électroniques (WEEE),
cet appareil ne peut pas être éliminé avec les déchets
domestiques.. Ceci est également valable pour les pays en
dehors de l’UE, selon leurs spécifications particulières.
S'il vous plaît de se débarrasser de ce produit conformément
à la réglementation locale, au point de collecte spécifié pour
les équipements électriques et électroniques.
En cas de questions, veuillez contacter l’autorité responsable
ou le distributeur auprès duquel cet appareil a été acheté.
Si cet appareil change de propriétaire (pour des raisons
personnelles ou professionnelles), cette consigne doit lui être
communiquée.
Pour des instructions de mise au rebut en Europe, voir
http://europe.ohaus.com/europe/en/home/support/weee.aspx.
Nous vous remercions de votre contribution à la protection de
l’environnement.
Frontier
TM
FC5306 FR-9
Remarque FCC
Cet équipement a été testé et déclaré conforme aux limites des appareils
numériques de classe B, selon la Section 15 des règles du FCC
(Commission fédérale des communications). Ces limites ont pour objectif
de fournir une protection raisonnable contre des interférences dangereuses
lorsque l'équipement est utilisé dans un environnement commercial. Cet
équipement génère, utilise et peut rayonner des hautes fréquences et s'il
n'est pas installé et utilisé conformément au guide de l'utilisateur, il peut
générer des interférences préjudiciables aux communications radio.
Cependant, il n'existe aucune garantie que des interférences ne se
produiront pas dans une installation particulière. Si cet équipement
provoque des interférences nuisibles à la réception radio ou télévision, ce
qui peut être déterminé en mettant l'équipement hors et sous tension,
l'utilisateur est encouragé à essayer de corriger l'interférence par une ou
plusieurs des mesures suivantes:
Réorienter ou déplacer l'antenne de réception.
Augmenter la distance entre l'équipement et le récepteur.
Connecter l'équipement à une sortie sur un circuit différent de celui
lequel le récepteur est connecté.
Consulter le revendeur ou un technicien radio / TV expérimenté.
Remarque d’Industrie Canada
Cet appareil numérique de classe B est conforme à la norme canadienne
ICES-001.
INHALTSVERZEICHNIS
1 EINLEITUNG ....................................................................................... 1
1.1 Sicherheitsmaßnahmen ....................................................................... 1
2 INSTALLATION ................................................................................... 3
2.1 Verpackungsinhalt ................................................................................ 3
2.2 Wechsel oder Installation des Rotors .................................................. 4
2.3 Einschalten ........................................................................................... 5
2.4 Vorbereitung für die Zentrifugierung .................................................... 5
2.5 Zentrifugierung (Betrieb) ...................................................................... 5
3 WICHTIGER HINWEIS ........................................................................ 6
3.1 Stromversorgung .................................................................................. 6
3.2 Verwendung und Wartung ................................................................... 6
4 TECHNISCHE DATEN ........................................................................ 7
5 EINHALTUNG ...................................................................................... 8
Frontier
TM
FC5306 DE-1
1 EINLEITUNG
Vielen Dank, dass Sie sich für dieses OHAUS-Produkt entschieden haben.
Bitte lesen Sie das Handbuch vollständig durch, bevor Sie die Frontier
TM
-
Zentrifuge FC5306 verwenden, um Fehler bei der Bedienung zu vermeiden.
Die Zentrifugen der Frontier
TM
5000-Serie sind mit vielen nützlichen
Funktionen ausgestattet. Dazu gehören beispielsweise:
Kleines und durchdachtes Gehäuse, Eintastenbedienung
Schnelle und einfache Installation sowie schneller Rotorwechsel
ohne Werkzeug
Der Rotor stoppt sofort, wenn der Deckel während des Betriebs
geöffnet wird.
Das Tischnetzteil ist für 100-240 V~ geeignet.
Ausgereiftes, geräuscharmes Produkt mit langen
Wartungsintervallen
1.1 Sicherheitsmaßnahmen
ACHTUNG: Lesen Sie zuerst alle Sicherheitshinweise, bevor
Sie das Gerät installieren, anschließen oder reparieren. Die
Nichtbeachtung der Warnhinweise kann zu Verletzungen
und/oder Beschädigungen führen. Bewahren Sie alle
Anweisungen zum späteren Nachschlagen auf.
Maßnahmen zu Ihrem Schutz
Arbeiten Sie nie in einer explosiven Umgebung! Das
Gehäuse des Geräts ist nicht gasdicht. (Es besteht
Explosionsgefahr aufgrund von Funkenbildung und
Korrosionsgefahr durch das Eindringen von Gasen)
Beachten Sie bei Verwendung von Chemikalien und
Lösungsmitteln die Hinweise des Herstellers sowie die
allgemeinen Laborsicherheitsvorschriften.
DE-2 Frontier
TM
FC5306
Maßnahmen für die Betriebssicherheit
Lehnen Sie sich nicht an die Zentrifuge.
Bleiben Sie nicht länger im Sicherheitsbereich der Zentrifuge als für die
Bedienung notwendig.
Lagern Sie keine potenziell gefährlichen Materialien innerhalb des
Sicherheitsbereichs der Zentrifuge.
Schrauben Sie die beiden Hälften des Gehäuses nicht ab.
Wischen Sie verschüttete Flüssigkeiten sofort auf! Das Gerät ist nicht
wasserdicht.
Prüfen Sie, ob die Eingangsspannung und der Netzstecker des
Netzteils zu der Netzspannung vor Ort passen.
Stellen Sie das Gerät so auf, dass Sie das Netzteil problemlos von der
Steckdose trennen können.
Das Netzkabel darf keine Stolpergefahr und kein Hindernis darstellen.
Das Gerät ist nur für die Verwendung in Räumen vorgesehen.
Verwenden Sie das Gerät nur an trockenen Standorten.
Verwenden Sie nur zugelassene Zubehörteile und Peripheriegeräte.
Verwenden Sie das Gerät nur unter den in der Anleitung angegebenen
Umgebungsbedingungen.
Trennen Sie die Waage vor dem Reinigen von der Stromversorgung.
Verwenden Sie die Waage nicht in gefährlichen oder instabilen
Umgebungen.
Die Waage darf nur von autorisierten Mitarbeitern gewartet und
repariert werden.
Vermeiden Sie die folgenden Umwelteinflüsse:
Starke Schwingungen
Direktes Sonnenlicht
Luftfeuchtigkeit über 80 %
Korrosive Gase
Temperaturen unter 5 °C und über 40 °C
Starke elektrische oder magnetische Felder
Frontier
TM
FC5306 DE-3
2 INSTALLATION
Nehmen Sie die Teile sorgfältig aus der Verpackung.
2.1 Verpackungsinhalt
Das Modell Frontier
TM
FC5306 umfasst folgende Teile:
Frontier
TM
FC5306
Foto
Einheiten
Frontier
TM
FC5306 -
Hauptgerät
1 Stück
Winkelrotor (8 x 1,5 / 2 ml)
1 Stück
4 x 0,2ml rechteckiger
Rotor für PCR 8-
Röhrchenreihe
1 Stück
Spezielles 24 V-Netzteil
1 Stück
Acht 0,5 ml-Adapter und
acht 0,2 ml-Adapter
8 x 2 Einheiten
Handbuch
1 Stück
DE-4 Frontier
TM
FC5306
2.2 Wechsel oder Installation des Rotors
1. Bei Lieferung vom Werk sind die Klemmhülse der Welle und der
Winkelrotor fest auf der Welle des Antriebsmotors montiert.
2. Um einen installierten Rotor auszubauen, müssen Sie die Außenkante
des Rotors festhalten, die hervorstehende Klemme mit beiden
Daumen in die Mitte drücken und dann den Rotor nach oben heben.
3. Um einen Rotor einzusetzen, müssen Sie die ovale Bohrung des
Rotors auf die Klemme ausrichten und dann den Rotor vorsichtig nach
unten drücken, bis die Klemme einrastet.
角转头
电机轴套
Motorwellenhülse
Winkelrotor
Frontier
TM
FC5306 DE-5
2.3 Einschalten
1. Sie können das Gerät mit dem Netzschalter auf der Rückseite
ausschalten.
2. Verbinden Sie das Ausgangsende des Netzteils mit dem
Netzeingang auf der Rückseite der Zentrifuge und schließen Sie
dann das Gerät an eine Stromquelle an.
2.4 Vorbereitung für die Zentrifugierung
1. Öffnen Sie den durchsichtigen Deckel (siehe folgendes Foto) und
setzen Sie die Proben symmetrisch ein (siehe Abschnitt “3.2
Verwendung und Wartung“)
2. Schließen Sie den Deckel und schalten Sie den Netzschalter auf der
Rückseite des Geräts ein (Hinweis: Der Rotor dreht sich nicht,
solange der Deckel geöffnet ist).
2.5 Zentrifugierung (Betrieb)
1 Sobald der Deckel geschlossen und das Gerät eingeschaltet ist,
beginnt der Rotor automatisch mit zunehmender Geschwindigkeit zu
drehen. Wenn die Nenngeschwindigkeit erreicht ist, dreht sich der
Rotor mit dieser Geschwindigkeit weiter.
2. Öffnen Sie den Deckel, wenn die erforderliche Zentrifugierung
abgeschlossen ist; der Rotor verlangsamt sich sofort, bis er
vollständig steht.
3. Sie müssen das Gerät nicht erst ausschalten, wenn eine weitere
Zentrifugierung notwendig ist.
4. Schalten Sie das Gerät aus und ziehen Sie den Netzstecker, wenn
die Zentrifugierungsarbeiten abgeschlossen sind oder das Gerät
nicht verwendet wird.
Nach unten
drücken,
um den Deckel
zu öffnen.
DE-6 Frontier
TM
FC5306
3 WICHTIGER HINWEIS
3.1 Stromversorgung
1. Die Stromversorgung des Geräts erfolgt über das Netzteil, das im
Lieferumfang des Geräts enthalten ist. Der Stecker des Netzteils darf
nicht mit einer dreiphasigen Steckdose verbunden werden.
2. Verbinden oder trennen Sie den Stecker des Netzteils nicht mit
nassen Händen. Ziehen Sie nicht am Kabel, um den Stecker
herauszuziehen.
3.2 Verwendung und Wartung
1. Bereiten Sie die Mikro-Bürette der Standardspezifikation mit dem
Klappdeckel auf die Zentrifugierung vor. Sie dürfen die Bürette nur bis
ca. 70 % der Nennkapazität mit Proben füllen. Prüfen Sie die
symmetrische Anordnung beim Einsetzen der Proben visuell. Achten
Sie darauf, dass kein giftiges oder infektiöses Probenmaterial austritt.
Achten Sie darauf, dass kein entzündliches oder explosives
chemisches Probenmaterial austritt.
2. Die Zentrifugenröhrchen mit dem Probenmaterial müssen
symmetrisch in die Röhrchenaussparungen des Zentrifugenrotors
platziert werden, damit der Rotor ausgewuchtet bleibt.
3. Vermeiden Sie beim Einsetzen oder Ersetzen des Rotors eine zu
starke Seitenkraft, damit der Rotor sich nicht verformt oder verbiegt.
Wenn die Klemmhülse um die Motorwelle etwas locker ist, muss die
M3-Stellschraube etwas nachgezogen werden.
4. Verwenden Sie keinen gebrochenen oder verformten Rotor.
5. Wischen Sie das Gerät bei der täglichen Reinigung und Wartung mit
sauberem Wasser oder einem neutralen Reinigungsmittel ab und
lassen Sie es an der Luft trocknen.
6. Achten Sie bei der Verwendung besonders auf die symmetrische und
horizontale Platzierung der Zentrifugenröhrchen.
VORSICHT: Trennen Sie das Netzteil vor der Reinigung von
der Steckdose und vom Gerät. Es dürfen keine Flüssigkeiten in
das Netzteil oder das Gerät eindringen.
Frontier
TM
FC5306 DE-7
4 TECHNISCHE DATEN
Die technischen Daten gelten unter folgenden Umgebungsbedingungen:
Betriebstemperatur: 5 °C bis 40 °C
Relative Luftfeuchtigkeit: 20 % bis 85 % relative Luftfeuchtigkeit, nicht
kondensierend
Höhe über Meeresspiegel: bis 2000 m
Verschmutzungsgrad: 2
Installationskategorie: II
Netzspannungsschwankungen: bis ± 10 % der Nennspannung
Modell
Frontier-MinizentrifugeFC5306
Rotor
Winkelrotor
hrchenreihenrotor
Max.
Rotationsgeschwindigkeit
6000 U/min ± 5%
Zentrifugalkraft
2000 g ± 10%
Rotationsrichtung
Gegen den Uhrzeigersinn
Max. Kapazität
8 x 1,5 ml / 2 ml
4 x 0,2 ml-PCR 8-
Röhrchenreihe
oder 32 x 0,2 ml einzelne
Zentrifugenröhrchen
Netzteil
0,2 ml / 0,5 ml
-
Beschleunigung
≤ 2 s
Abbremsung
≤ 3 s
Antriebsmodus und Motor
Direktantrieb, geräuscharm dank Rollenlager
Stromversorgung
Eingang: 100–240 V~, 50/60 Hz, 1,5 A
Ausgang: 24 V–
Abmessungen (B × T × H)
5,9 × 6,6 × 5,0 in. / 150 × 168 × 128 mm
Nettogewicht
2,316 lb / 1,05 kg (mit Freilast-Winkelrotor)
Versandabmessungen
(B × T × H)
8,5 × 12,0 × 7,6 in. / 216 × 305 × 194 mm
Versandgewicht
4,079 lb / 1,85 kg
.6A
DE-8 Frontier
TM
FC5306
5 EINHALTUNG
Die Einhaltung folgender Normen ist durch eine entsprechende
Kennzeichnung auf dem Produkt vermerkt.
Kenn-
zeichnung
Norm
Dieses Produkt entspricht der EMV-Richtlinie 2004/108/EG und
der Niederspannungsrichtlinie 2006/95/EG. Die vollständige
Konformitätserklärung finden Sie online unter
http://europe.ohaus.com/europe/en/home/support/compliance/c
e-declaration-of-conformity.aspx
Entsorgung
Entsprechend der Europäischen Richtlinie 2002/96/EG über
Elektro- und Elektronik-Altgeräte (WEEE) darf dieses Gerät
nicht im Hausmüll entsorgt werden. Dies gilt auch für Länder
außerhalb der EU entsprechend den jeweiligen Vorschriften.
Entsorgen Sie dieses Produkt entsprechend den lokalen
Vorschriften an der Sammelstelle für Elektro- und
Elektronikgeräte.
Bei Fragen wenden Sie sich bitte an die entsprechende
Behörde oder an den Händler, von dem Sie das Gerät erworben
haben.
Wenn Sie das Gerät an andere weitergeben (zur privaten oder
beruflichen Nutzung), muss der Inhalt dieser Vorschriften
ebenfalls weitergeben werden.
Entsorgungshinweise für Europa finden Sie unter
http://europe.ohaus.com/europe/en/home/support/weee.aspx.
Vielen Dank für Ihren Beitrag zum Umweltschutz.
Frontier
TM
FC5306 DE-9
FCC-Hinweis
Jedes Gerät wurde geprüft und entspricht den Grenzwerten für
Digitalgeräte der Klasse B gemäß Abschnitt 15 der FCC-Vorschriften.
Diese Grenzwerte sollen in Wohnumgebungen einen angemessenen
Schutz gegen Funkstörungen gewährleisten. Dieses Gerät erzeugt und
verwendet HF-Energie und strahlt diese u. U. auch ab; wird es nicht
entsprechend der Anweisung installiert und verwendet, können
Funkstörungen auftreten. Es gibt jedoch keine Garantie, dass bei einer
bestimmten Installation keine Störung auftritt. Wenn dieses Gerät zu
Funkstörungen beim Radio- oder Fernsehempfang führt, was sich durch
Ein- und Ausschalten des Geräts überprüfen lässt, sollte der Benutzer die
Störung durch eine oder mehrere der folgenden Maßnahmen beheben:
Neuausrichtung oder Umsetzung der Empfangsantenne
Größerer Abstand zwischen Gerät und Empfänger
Anschluss des Geräts an eine Steckdose in einem anderen Stromkreis
als der Empfänger
Beratung durch den Händler oder einen erfahrenen Radio-
/Fernsehtechniker
Industry Canada-Hinweis
Das Digitalgerät der Klasse B entspricht den kanadischen Vorschriften
ICES-001.
INDICE
1 INTRODUZIONE .......................................................... 1
1.1 Misure di sicurezza .............................................................................. 1
2 INSTALLAZIONE ........................................................ 3
2.1 Contenuto della confezione ................................................................. 3
2.2 Sostituzione o installazione del rotore ................................................. 4
2.3 Accensione ........................................................................................... 5
2.4 Preparazione per la centrifuga ............................................................. 5
2.5 Centrifuga (In esecuzione) ................................................................... 5
3 ATTENZIONE SPECIALE ......................................... 6
3.1 Alimentazione ....................................................................................... 6
3.2 Uso e manutenzione ............................................................................ 6
4 SPECIFICHE TECNICHE .......................................... 7
5 CONFORMITÀ ............................................................. 8
Frontier
TM
FC5306 EN-1
1 INTRODUZIONE
Grazie per aver scelto questo prodotto OHAUS.
Prima di usare Frontier™ FC5306 leggere il manuale interamente per
evitare operazioni non corrette.
La centrifuga della serie Frontier
TM
5000 include molte funzioni utili. Alcune
tra queste funzioni per Frontier
TM
FC5306 sono:
Design piccolo e molto bello, operatività con un solo tasto
Installazione conveniente e rapida e sostituzione del rotore senza
attrezzi
Il movimento del rotore si interrompe immediatamente se durante il
funzionamento viene aperto il coperchio
L'adattatore di alimentazione del desktop è applicabile a
alimentazioni da 100-240 V CA
Prodotto elaborato poco rumoroso e con vita utile lunga
1.1 Misure di sicurezza
ATTENZIONE: leggere tutte le avvertenze di sicurezza prima di
procedere all'installazione, alla realizzazione dei collegamenti o
alla manutenzione dell'apparecchiatura. Il mancato rispetto delle
avvertenze suddette può provocare lesioni personali e/o danni
alla proprietà. Conservare tutte le istruzioni per poterle
consultare in caso di necessità.
Misure di protezione personale
Non lavorare mai in un ambiente soggetto a pericolo di
esplosione! L'alloggiamento dello strumento non è a prova
di gas. (pericolo di esplosione dovuto a formazione di
scintille, corrosione causata da ingresso di gas)
Quando si utilizzano prodotti chimici e solventi, seguire le
istruzioni del produttore e le regole di sicurezza presenti
sull'etichetta generica.
Frontier
TM
FC5306 IT-1
IT-2 Frontier
TM
FC5306
Misure per la sicurezza operativa
Non appoggiarsi sulla centrifuga.
Non rimanere all'interno dell'area di sicurezza più a lungo del
necessario per motivi operativi.
Non depositare materiali potenzialmente pericolosi all'interno dell'area
di sicurezza.
Non svitare le metà dell'alloggiamento.
Asciugare immediatamente eventuali liquidi versati! Lo strumento non è
impermeabile.
Verificare che l’intervallo di tensione in ingresso dell'adattatore CA e il
tipo di spina siano adatti all'alimentazione elettrica locale.
Non posizionare l'apparecchiatura in modo tale che risulti difficoltoso
scollegare l'adattatore CA dalla presa elettrica.
Accertarsi che il cavo di alimentazione non rappresenti un potenziale
ostacolo o non costituisca un intralcio al passaggio.
Quest'apparecchiatura è destinata esclusivamente all'uso interno.
Utilizzare l'apparecchio solo in luoghi asciutti.
Utilizzare solo accessori e periferiche approvate.
Utilizzare l’apparecchiatura solo nelle condizioni ambientali specificate
in queste istruzioni.
Prima di effettuare la pulizia dell’apparecchiatura, scollegare
l'alimentazione elettrica.
Non utilizzare l’apparecchiatura in ambienti instabili o pericolosi.
La manutenzione deve essere eseguita soltanto da personale
autorizzato.
Escludere le seguenti influenze ambientali:
Vibrazioni potenti
Luce solare diretta
Umidità atmosferica superiore all'80%
Presenza di gas corrosive
Temperature inferiori a 5 °C e superiori a 40 °C
Campi elettrici o magnetici potenti
Frontier
TM
FC5306 IT-3
2 INSTALLAZIONE
Disimballaggio con cura.
2.1 Contenuto della confezione
Il modello Frontier
TM
FC5306 presenta i seguenti elementi:
Frontier
TM
FC5306
Foto
Unità
Macchina principale
Frontier
TM
FC5306
1 set
Rotore angolare
(8 x 1,5 / 2ml)
1 set
Rotore rettangolare
4 x 0,2ml per banco a
8 provette in PCR
1 set
Alimentatore dedicato
CC 24 V
1 set
Otto adattatori da 0,5 ml
e otto adattatori da 0,2 ml
8 x 2 unità
Manuale
1 set
IT-4 Frontier
TM
FC5306
2.2 Sostituzione o installazione del rotore
1. Quando la macchina viene consegnata dalla fabbrica, il manicotto
dell'albero di blocco e il rotore angolare sono fermi in modo fisso
sull'albero del motore a trazione.
2. Se è necessario smontare un rotore installato, l'utente deve
mantenere il bordo esterno del rotore e spingere il morsetto che
protrude verso il centro con i pollici, quindi sollevare in alto il rotore.
3. Per reinstallare il rotore, allineare il foro ovale del rotore sopra al
morsetto, quindi premere delicatamente verso il basso il rotore fino a
quando il morsetto non scatta in posizione.
角转头
电机轴套
Rotore angolare
Manicotto albero
motore
Morsetto
Frontier
TM
FC5306 IT-5
2.3 Accensione
1. Con l'interruttore di alimentazione sul retro della macchina su OFF.
2. Collegare l'estremità di uscita dell'adattatore nella presa sul retro
della macchina e quindi collegare a una presa elettrica.
2.4 Preparazione per la centrifuga
1. Aprire il coperchio trasparente (vedere foto sotto) e aggiungere in
modo corretto il carico e bilanciarlo (fare riferimento alla sezione
“3.2 uso e manutenzione”). (refer to “section 3.2 use and maintenance”).
2. Chiudere il coperchio e premere l'interruttore di alimentazione sul
retro della macchina in posizione ON (nota: il rotore non funziona
con il coperchio aperto).
2.5 Centrifuga (In esecuzione)
1 Dopo aver chiuso il coperchio a acceso la macchina, il rotore inizia
automaticamente a funzionare aumentando la velocità fino alla
velocità nominale e quindi funziona in modo stabile.
2. Aprire il coperchio quando sono soddisfatti i requisiti di
centrifugazione, il rotore rallenta immediatamente fino all'arresto
completo.
3. Non è necessario spegnere se si deve fare un'altra centrifugazione.
4. Spegnere e scollegare la macchina dalla presa elettrica al termine
del lavoro di centrifugazione o se non è in uso.
Premere verso il
basso per aprire
il coperchio
IT-6 Frontier
TM
FC5306
3 ATTENZIONE SPECIALE
3.1 Alimentazione
1. Questa macchina deve essere alimentata tramite l'adattatore fornito.
La spina CA dell'adattatore non deve essere collegata a fonti elettriche
trifase.
2. Non inserire la spina o l'adattatore con le mani bagnate.
Non tirare con forza il cavo per staccare la spina.
3.2 Uso e manutenzione
1. Preparare la micro buretta con le specifiche standard con il coperchio
con cerniera per la centrifugazione. È possibile aggiungere campione
fino a circa il 70% della capacità nominale al massimo. Controllare
visivamente il bilanciamento quando si aggiunge il campione. Non
staccare campioni biologici velenosi o infetti. Non staccare campioni
chimici infiammabili o velenosi.
2. Le provette per la centrifuga con campione devono essere posizionate
in modo simmetrico nei fori per provetta per centrifuga del rotore per
mantenere il rotore bilanciato.
3. Quando viene installato o sostituito il rotore, evitare un'eccessiva forza
laterale per impedire deformazioni o piegature. Se il manicotto
dell'albero di blocco attorno all'albero del motore è un poco lento, è
necessario serrare di nuovo la vite M3.
4. Non utilizzare un rotore rotto o deformato.
5. Durante la pulizia e la manutenzione giornaliere, la macchina deve
essere pulita con acqua pulita o detergente neutro e asciugata tramite
ventilazione naturale.
6. Durante l'uso, fare attenzione a posizionare in modo simmetrico e
orizzontale le provette per la centrifuga.
AVVERTENZA: Prima di effettuare la pulizia
dell’apparecchiatura, scollegare l'adattatore Ca
dall'alimentazione elettrica CA. Non lasciare che liquidi
penetrino nell'adattatore CA o nell'apparecchio.
Frontier
TM
FC5306 IT-7
4 SPECIFICHE TECNICHE
I dati tecnici sono validi per le seguenti condizioni ambientali:
Temperatura di funzionamento: da 5 a 40
Umidità relativa: 20-85% di umidità relativa, senza condensa
Altitudine s.l.m.: Fino a 2.000 m
Grado di inquinamento: 2
Categoria d'installazione: II
Fluttuazioni di tensione di alimentazione minima: fino a ±10% della
tensione nominale
Modello
Centrifuga Frontier MiniFC5306
Rotore
Rotore angolare
Rotore con banco
provette
Velocità di rotazione
6000 giri/min ± 5%
Forza centrifuga
2000 g ± 10%
Direzione di rotazione
Senso antiorario
Capacità Max
8 x 1.5 ml / 2 ml
banco per 8 provette 4 x
0,2 ml PCR
o per provetta centrifuga
singola 32 x 0,2 ml
Adattatore
0,2 ml/0,5 ml
-
Tempo di accelerazione
≤ 2 s
Tempo di decelerazione
≤ 3 s
Modalità trasmissione e
motore
Trasmissione diretta poco rumore dovuto alle sfere rotanti
Alimentazione
Ingresso: 100-240 V ~ 50/60Hz 1.5A
Uscita: 24V
Dimensioni
(L x A x P)
150 × 168 × 128 mm / 5.9 × 6,6 × 5,0"
Peso netto
1,05 kg / 2.316 lb (incluso rotore angolo di caricamento
libero)
Dimensioni dell’imballo
(L × A × P)
216 × 305 × 194 mm / 8,5 × 12.0 × 7,6"
Peso di spedizione
1,85 kg / 4.,79 lb
.6A
IT-8 Frontier
TM
FC5306
5 CONFORMITÀ
La conformità agli standard seguenti viene indicata dalla presenza del
marchio corrispondente sul prodotto.
Marchio
Standard
Questo prodotto è conforme alla direttiva EMC 2004/108/CEE e
alla Direttiva Bassa Tensione 2006/95/CE. La Dichiarazione di
Conformità è disponibile online all'indirizzo
http://europe.ohaus.com/europe/en/home/support/compliance/c
e-declaration-of-conformity.aspx
Smaltimento
In conformità alla Direttiva Europea 2002/96/CE sullo
smaltimento delle apparecchiature elettriche ed elettroniche
(RAEE), l’apparecchiatura non deve essere smaltita assieme ad
altri rifiuti domestici. Queste disposizioni sono valide anche nei
paesi esterni all'UE, in base ai requisiti delle varie legislazioni.
Smaltire questo prodotto secondo le normative locali, presso il
punto di raccolta specificato per le apparecchiature elettriche ed
elettroniche.
In caso di dubbi, rivolgersi all'ente responsabile o al distributore
da cui è stato acquistato questo dispositivo.
Qualora l’apparecchiatura sia trasferita a terzi (per uso privato o
professionale), il contenuto della presente normativa deve
ritenersi valido anche per terzi.
Per le istruzioni relative allo smaltimento in Europa, fare
riferimento a
http://europe.ohaus.com/europe/en/home/support/weee.aspx.
Grazie per il contributo alla salvaguardia
dell'ambiente.conformance.
Frontier
TM
FC5306 IT-9
Nota per FCC
Questa apparecchiatura è stata collaudata e ritenuta conforme ai limiti dei
dispositivi digitali di Classe B, in conformità con la Parte 15 delle norme
FCC. Tali limiti sono previsti per fornire un’adeguata protezione contro
interferenze dannose durante il funzionamento dell'apparecchiatura in
un'installazione residenziale. L’apparecchiatura genera, utilizza e p
irradiare energia nella banda delle radiofrequenze, per cui, se non installata
e utilizzata in conformità alle istruzioni, potrebbe causare interferenze
dannose con le comunicazioni radio. Tuttavia, non vi è garanzia che
l'interferenza non si verifichi in una particolare installazione. Se tale
apparecchiatura causa realmente interferenza pericolosa alla ricezione di
radio e televisione, determinabile mediante lo spegnimento e la
riaccensione dell'apparecchiatura, l'utente è incoraggiato a provare a
correggere l'interferenza adottando una delle seguenti misure:
Riorientare o mettere in un altro posto l'antenna di ricezione.
• Aumentare la distanza di separazione tra l'apparecchio e il ricevitore.
• Collegare l'apparecchio a una presa su un circuito diverso da quello cui
è collegato il ricevitore.
• Consultare il rivenditore o un tecnico radio/TV esperto per aiuto.
Nota per il Canada
Questa apparecchiatura digitale di classe B è conforme alla normativa
canadese ICES-001.
Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054 USA
With offices worldwide / Con oficinas en todo el mundo / Avec des bureaux
dans le monde entier / Mit weltweiten Niederlassungen / Con uffici in tutto il
mondo
http://www.ohaus.com
*30129575*
P/N 30129575 G © 2020 OHAUS Corporation, all rights reserved / todos
los derechos reservados / tous droits réservés / Alle Rechte vorbehalten /
tutti i diritti riservati
Printed in China / Impreso en China / Imprimé en Chine / Gedruckt in China
/ Stampato in Cina

Transcripción de documentos

FrontierTM Centrifuge FC5306 Instruction Manual Centrifuga FrontierTM FC5306 Manual de instrucciones Centrifugeuse FrontierTM FC5306 Manuel d’instructions FrontierTM-Zentrifuge FC5306 Bedienungsanleitung Centrifuga FrontierTM FC5306 Manuale di istruzioni TABLE OF CONTENTS 1 INTRODUCTION ..............................................................1 1.1 Safety measures .................................................................................. 1 2 INSTALLATION................................................................3 2.1 Package contents ................................................................................. 3 2.2 Replacement or Installation of Rotor .................................................... 4 2.3 Power On ............................................................................................. 5 2.4 Preparation for Centrifugation .............................................................. 5 2.5 Centrifugation (Running) ...................................................................... 5 3 SPECIAL ATTENTION .....................................................6 3.1 Power supply ........................................................................................ 6 3.2 Use and maintenance .......................................................................... 6 4 SPECIFICATIONS ............................................................7 5 COMPLIANCE ..................................................................8 FrontierTM FC5306 1 EN-1 INTRODUCTION Thank you for choosing this OHAUS product. TM Please read the manual completely before using the Frontier FC5306 to avoid incorrect operation. TM You will find much helpful functionality in Frontier 5000 series centrifuge. TM A few of the many existing features for Frontier FC5306 are:  Small and exquisite body, one-key operation  Convenient and quick installation and replacement of rotor without    1.1 need of using tool Rotor will immediately stop running if lid is opened during operation Desktop power adapter is applicable to 100-240V AC power supply Elaborate product with low noise and long service time Safety measures CAUTION: Read all safety warnings before installing, making connections, or servicing this equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain all instructions for future reference. Measures for your protection  Never work in an environment subject to explosion hazards! The housing of the instrument is not gas tight. (explosion hazard due to spark formation, corrosion caused by the ingress of gases)  When using chemicals and solvents, comply with the instructions of the producer and the general lab safety rules. Measures for operational safety  Do not lean on the centrifuge.  Do not stay within the clearance envelope longer than necessary for operational reasons.  Do not deposit any potentially hazardous materials within the clearance envelope.  Do not unscrew the two halves of the housing.  Dry off any liquid spills immediately! The instrument is not watertight. FrontierTM FC5306 EN-2          Verify that the AC adapter’s input voltage range and plug type are compatible with the local AC power to be used. Do not position the equipment such that it is difficult to disconnect the AC adapter from the power receptacle. Make sure that the power cord does not pose a potential obstacle or tripping hazard. The equipment is for indoor use only. Use the equipment only in dry locations. Use only approved accessories and peripherals. Operate the equipment only under ambient conditions specified in these instructions. Disconnect the equipment from the power supply when cleaning. Do not operate the equipment in hazardous or unstable environments. Service should only be performed by authorized personnel. Exclude the following environmental influences:       Powerful vibrations Direct sunlight Atmospheric humidity greater than 80% Corrosive gases present Temperatures below 5 °C and above 40 °C Powerful electric or magnetic fields FrontierTM FC5306 2 EN-3 INSTALLATION Carefully unpack the box. 2.1 Package contents The model Frontier Frontier TM TM FC5306 has the following items: FC5306 Photo Units TM Frontier FC5306 main machine 1 set Angle rotor ( 8 x 1.5/2ml) 1 set 4 x 0.2ml rectangle rotor for PCR 8-tube bank 1 set Dedicated DC24V power adapter 1 set Eight 0.5 ml adapters and eight 0.2 ml adapters 8 x 2 units Manual 1 set FrontierTM FC5306 EN-4 2.2 1. Replacement or Installation of Rotor When the machine is delivered from the factory, the seizing shaft sleeve and angle rotor are firmly fixed on the shaft of the drive motor. 2. If it is necessary to dismantle an installed rotor, the user should hold the outside edge of the rotor, and push the protruding clamp towards its center with two thumbs, then lift the rotor upwards. 3. To reinstall rotor, align the oval hole of rotor over the clamp, then gently press down the rotor until the clamp clicks into place. 角转头 Angle Rotor Clamp Motor Shaft Sleeve 电机轴套 FrontierTM FC5306 2.3 1. 2. 2.4 1. 2. 2.5 1. 2. 3. 4. EN-5 Power On With the power switch on the back of this machine OFF. Connect the output end of the adapter to the power socket on the back of the machine, and then plug the unit to a power source. Preparation for Centrifugation Open the transparent lid (see photo below) and properly add load in a balanced manner (refer to section “3.2 use and maintenance”). Close the lid then press the power switch on the back of this machine to ON (note: The rotor will not run as long as the lid is open). Centrifugation (Running) After the lid is closed and the machine switched on, the rotor will automatically start running with its speed increasing to the rated speed, and then it will run stably. Open the lid when the requirement of centrifugation is met, the rotor will immediately slow down to a complete stop. It is unnecessary to turn off power if another centrifugation is required. Turn off and unplug the machine when all centrifugation work is completed or when not in use. press down to open the lid FrontierTM FC5306 EN-6 3 3.1 1. 2. 3.2 1. 2. 3. 4. 5. 6. SPECIAL ATTENTION Power supply The power supply to this machine must be realized through the adapter provided together with the machine. The AC plug of adapter should not be connected with three-phase power source. Don’t insert or disconnect the power plug of the adapter with wet hand. Don’t forcibly pull the cable to disconnect the power plug. Use and maintenance Prepare the micro burette of standard specification with hinged cover for centrifugation. It is only allowed to add sample till approx. 70% of nominal capacity at most. Visually check balance during adding of sample. Don’t detach any poisonous or infectious biological sample. Don’t detach flammable or explosive chemical sample. The centrifugal tubes with sample must be symmetrically placed into the centrifugal tube holes of the rotor to keep the rotor balanced. When the rotor is installed or replaced, excessive sidewise force should be avoided to prevent its deformation and bending. If the seizing shaft sleeve around the motor shaft is a bit loose, it is necessary to fasten the M3 set screw again. Do not use broken or deformed rotor. During daily cleaning and maintenance, the machine should be wiped with clean water or neutral detergent, and be dried through natural ventilation. During use, attention should be paid to symmetrical and horizontal placement of centrifugal tubes. WARNING: Disconnect the AC adapter from the AC power supply and the equipment before cleaning. Do not allow liquids to enter the AC adapter or the equipment. FrontierTM FC5306 4 EN-7 SPECIFICATIONS The technical data is valid under the following ambient conditions: Operating temperature: 5°C to 40°C Relative humidity: 20% to 85% relative humidity, non-condensing Height above sea level: Up to 2000 m Pollution degree: 2 Installation category: II Main supply voltage fluctuations: up to ± 10% of the nominal voltage Model Frontier Mini Centrifuge:FC5306 Rotor Angle Rotor Tube Bank Rotor Rotating Speed Scope 6000 rpm ± 5% Centrifugal Force 2000 g ± 10% Rotating Direction Counter clockwise Max Capacity 8 x 1.5 ml / 2 ml 4 x 0.2 ml-PCR 8-tube bank or 32 x 0.2 ml single centrifugal tube Adapter 0.2 ml / 0.5 ml - Accelerating Time ≤2s Decelerating Time ≤3s Drive Mode and Motor Direct drive, with low noise attributable to rolling bearing Power Supply Input: 100-240V ~ 50/60Hz 1.5A Output: 24V .6 A Dimensions (W × D × H) 5.9 × 6.6 × 5.0 in. / 150 × 168 × 128 mm Net Weight 2.316 lb / 1.05 Kg (including free-load angle rotor) Shipping Dimensions (W × D × H) 8.5 × 12.0 × 7.6 in. / 216 × 305 × 194 mm Shipping Weight 4.079 lb / 1.85 Kg FrontierTM FC5306 EN-8 5 COMPLIANCE Compliance to the following standards is indicated by the corresponding mark on the product. Marking Standard This product conforms to the EMC directive 2004/108/EC and the Low Voltage Directive 2006/95/EC. The complete Declaration of Conformity is available online at http://europe.ohaus.com/europe/en/home/support/compliance/c e-declaration-of-conformity.aspx Disposal In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. For disposal instructions in Europe, refer to http://europe.ohaus.com/europe/en/home/support/weee.aspx. Thank you for your contribution to environmental protection. FrontierTM FC5306 EN-9 FCC Note This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: • Reorient or relocate the receiving antenna. • Increase the separation between the equipment and receiver. • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. • Consult the dealer or an experienced radio/TV technician for help. Industry Canada Note This Class B digital apparatus complies with Canadian ICES-001. TABLA DE CONTENIDO 1 INTRODUCCIÓN ..............................................................1 1.1 Medidas de seguridad .......................................................................... 1 2 INSTALACIÓN .................................................................3 2.1 Contenido del paquete ......................................................................... 3 2.2 Reemplazo o instalación del rotor ........................................................ 4 2.3 Encendido ............................................................................................ 5 2.4 Preparación para la centrifugación ...................................................... 5 2.5 Centrifugación (operación) ................................................................... 5 3 ATENCIÓN ESPECIAL ....................................................6 3.1 Alimentación eléctrica .......................................................................... 6 3.2 Uso y mantenimiento ........................................................................... 6 4 ESPECIFICACIONES .......................................................7 5 CONFORMIDAD NORMATIVA ........................................8 FrontierTM FC5306 1 ES-1 INTRODUCCIÓN Gracias por elegir este producto OHAUS. TM Lea este manual completamente antes de usar la Frontier FC5306 para evitar la operación incorrecta. TM Encontrará mucha funcionalidad útil en la centrífuga Frontier serie 5000. TM Algunas de las muchas características actuales de la Frontier FC5306 son:  Cuerpo pequeño y refinado, y operación con una tecla  Instalación conveniente y rápida y reemplazo del rotor sin necesidad de usar herramientas  El rotor se detiene inmediatamente si se abre la tapa durante la operación  El adaptador de corriente de mesa es compatible con la corriente eléctrica de 100 a 240 V  Bajo nivel de ruido durante la operación. 1.1 Medidas de seguridad PRECAUCIÓN: Lea todas las advertencias de seguridad antes de instalar, hacer conexiones o dar servicio a este equipo. La falta de cumplimiento de estas advertencias puede resultar en lesiones personales o daños materiales. Conserve todas las instrucciones para referencia futura. Medidas para su protección  Nunca trabaje en un ambiente sujeto a peligros de explosión El equipo no es impermeable al gas. (Peligro de explosión debido a formación de chispas; corrosión debido al ingreso de gases).  Cuando use químicos y solventes, siga las instrucciones del fabricante y reglas de seguridad generales de laboratorio. ES-2 FrontierTM FC5306 Medidas para seguridad operativa  No se apoye sobre la centrífuga.  No permanezca dentro del espacio libre de seguridad más tiempo del necesario por razones operativas.  No deposite ninguna sustancia peligrosa dentro del espacio libre de seguridad.  No destornille la tapa del equipo.  Seque de inmediato cualquier líquido que se derrame. El instrumento no es hermético al agua.  Verifique que el rango del voltaje de entrada del adaptador de corriente y el tipo de enchufe sean compatibles con la corriente alterna local que va a usarse.  No coloque el equipo de tal forma que sea difícil desconectar el adaptador de corriente de la toma de corriente.  Verifique que el cordón eléctrico no represente un obstáculo potencial o riesgo de tropezarse.  Este equipo es solamente para uso en interiores. Use el equipo solamente en lugares secos.  Use solo accesorios y periféricos aprobados.  Opere el equipo solo en las condiciones ambientales especificadas en estas instrucciones.  Desconecte el equipo de la alimentación eléctrica antes de limpiarlo.  No opere el equipo en ambientes peligrosos o inestables.  El servicio debe proporcionarse solamente por personal autorizado. Excluya las siguientes influencias del medio ambiente:  Vibraciones potentes  Luz solar directa  Humedad atmosférica mayor del 80%  Presencia de gases corrosivos  Temperaturas menores de 5 °C y mayores de 40 °C  Campos eléctricos o magnéticos potentes FrontierTM FC5306 2 ES-3 INSTALACIÓN Desempaque con cuidado la caja. 2.1 Contenido del paquete El modelo Frontier Frontier TM TM FC5306 tiene los siguientes componentes: FC5306 Imagen Unidades TM Frontier FC5306 Equipo principal 1 Rotor angular (8 x 1.5/2 ml) 1 Rotor rectangular 4 x 0.2 ml para banco PCR de 8 tubos 1 Adaptador de corriente de 24 VCD 1 Ocho adaptadores de 0.5 ml y ocho adaptadores de 0.2 ml Manual 8 x 2 unidades 1 FrontierTM FC5306 ES-4 2.2 1. Reemplazo o instalación del rotor Cuando el equipo se entrega de fábrica, el sujetador y el rotor angular se fijan firmemente en el eje del motor. 2. Si es necesario reemplazar un rotor instalado, el usuario deberá sostener el borde externo del rotor y empujar la abrazadera sobresaliente hacia el centro con los dos pulgares, y luego mover el rotor hacia arriba. 3. Para instalar el rotor, alinie el orificio ovalado del rotor sobre la abrazadera y luego presione hacia abajo el rotor hasta que la abrazadera haga clic en su lugar. Rotor angular 角转头 Abrazadera Sujetador 电机轴套 FrontierTM FC5306 2.3 1. 2. 2.4 1. 2. 2.5 1. 2. 3. 4. ES-5 Encendido El interruptor de corriente en la parte posterior del equipo debe estar en OFF (Apagado). Conecte el extremo de salida del adaptador en el enchufe de corriente de la parte posterior del equipo, y luego conecte la unidad en una fuente de energía. Preparación para la centrifugación Abra la tapa transparente (vea la foto siguiente) y agregue carga apropiadamente en una forma balanceada (consulte la sección “3.2 Uso y mantenimiento”). Cierre la tapa y luego presione el interruptor de corriente en la parte posterior de la máquina a la posición ON (Encendido) (Nota: El rotor no funcionará mientras la tapa esté abierta). Centrifugación (operación) Después de cerrar la tapa y de encender el equipo, el rotor comenzará a funcionar automáticamente y la velocidad se incrementará hasta la velocidad nominal, y luego funcionará de manera estable. Abra la tapa cuando se cumpla el requerimiento de centrifugación, y el rotor reducirá su velocidad inmediatamente hasta detenerse por completo. Para realizar otra centrofugación, solo cierre la tapa nuevamente. Apague y desenchufe la máquina cuando haya terminado todo el trabajo de centrifugación o cuando no esté en uso Presione para abrir la tapa FrontierTM FC5306 ES-6 3 ATENCIÓN ESPECIAL 3.1 Alimentación eléctrica 1. 2. 3.2 1. 2. 3. 4. 5. 6. La alimentación eléctrica hacia este equipo debe hacerse a través del adaptador que se proporciona con el equipo. El enchufe de corriente del adaptador no deberá conectarse con la alimentación eléctrica trifásica. No inserte ni desconecte el enchufe de corriente del adaptador con las manos húmedas. No jale el cable con fuerza para desconectar el enchufe de corriente. Uso y mantenimiento Prepare la microbureta de especificación estándar con cubierta de bisagras para centrifugación. Solo se permite agregar muestra Un 70% de la capacidad nominal. No ingrese ninguna muestra biológica venenosa o infecciosa. No ingrese ninguna muestra química inflamable o explosiva. Los tubos centrífugos con muestra deben colocarse en forma simétrica en los orificios del rotor para mantener el rotor balanceado. Al instalar o reemplazar el rotor, se deberá evitar ejercer fuerza lateral excesiva para prevenir la deformación y flexión. Si el sujetador de agarre alrededor del eje del motor está un poco flojo, es necesario apretar nuevamente el tornillo de sujeción M3. No use el rotor si está roto o deformado. Durante la limpieza y mantenimiento diario, el equipo debe frotarse con agua o detergente neutro, y debe secarse mediante ventilación natural. Durante el uso, se debe prestar atención a la colocación simétrica y horizontal de los tubos centrífugos. ADVERTENCIA: Desconecte el adaptador de corriente de la alimentación eléctrica y del equipo antes de la limpieza. No permita que entren líquidos al adaptador de corriente ni al equipo. FrontierTM FC5306 4 ES-7 ESPECIFICACIONES Los datos técnicos son válidos en las siguientes condiciones ambientales: Temperatura de operación: 5 °C a 40 °C Humedad relativa: humedad relativa de 20% a 85%, no condensante Altura sobre el nivel del mar: hasta 2000 m Grado de contaminación: 2 Categoría de instalación: II Fluctuaciones de voltaje de la alimentación principal: hasta ± 10% del voltaje nominal Modelo Rotor Minicentrífuga Frontier:FC5306 Rotor angular Rotor de tubos Alcance de la velocidad de rotación 6000 rpm ± 5% Fuerza centrífuga 2000 g ± 10% Dirección de rotación Máxima capacidad Adaptador En sentido contrario a las manecillas del reloj 8 x 1.5 ml / 2 ml 0.2 ml / 0.5 ml 4 x 0.2 ml - Banco de 8 tubos PCR o tubo centrífugo simple de 32 x 0.2 ml - Tiempo de aceleración ≤2s Tiempo de desaceleración ≤3s Accionamiento del motor Alimentación eléctrica Dimensiones Peso neto Dimensiones del envío Peso de envío Accionamiento directo Entrada: 100-240 V ~ 50/60Hz 1.5 A Salida: 24 V .6 A 150 mm An x 168 mm Pr x 128 mm Al 5.9 pulg. An x 6.6 pulg. Pr x 5.0 pulg. Al 1.05 kg (incluyendo rotor angular) 2.316 lb 8.5 × 12.0 × 7.6 in. / 216 × 305 × 194 mm 4.079 lb / 1.85 Kg ES-8 5 FrontierTM FC5306 CONFORMIDAD NORMATIVA La marca correspondiente en el producto indica la conformidad con los siguientes estándares. Marca Estándar Este producto está en conformidad con la directiva EMC 2004/108/EC y con la directiva para Voltaje Bajo 2006/95/EC. La Declaración de conformidad completa está disponible en línea en http://europe.ohaus.com/europe/en/home/support/compliance/c e-declaration-of-conformity.aspx Disposición final De conformidad con la Directiva Europea 2002/96/EC sobre Residuos de Equipos Eléctricos y Electrónicos (WEEE), este dispositivo no puede desecharse con la basura doméstica. Esto también es aplicable para países fuera de la UE, según sus requerimientos específicos. Deseche este producto de acuerdo con las regulaciones locales en el punto de recolección especificado para equipos eléctricos y electrónicos. Si tiene alguna pregunta, comuníquese con la autoridad responsable o con el distribuidor a quien compró este dispositivo. En caso de que este dispositivo sea transferido a otras partes (para uso privado o profesional), también deberá mencionarse el contenido de esta regulación. Para consultar las instrucciones de eliminación en Europa, visite http://europe.ohaus.com/europe/en/home/support/weee.aspx. Gracias por su contribución a la protección ambiental. FrontierTM FC5306 ES-9 Nota de la FCC Este equipo ha sido probado y se encontró que cumple con los límites para un dispositivo digital clase B, consecuente con la Parte 15 de las Reglas de la FCC. Estos límites están diseñados para proporcionar una protección razonable contra interferencia dañina cuando el equipo es operado en un ambiente comercial. Este equipo genera, usa y puede irradiar frecuencias de radio y, si no es instalado y utilizado de acuerdo con el manual de instrucciones, puede causar interferencia dañina a las radiocomunicaciones. Sin embargo, no hay garantía de que no se produzcan interferencias en una instalación en particular. Si este equipo causa interferencias perjudiciales en la radio o televisión, lo cual puede comprobarse encendiéndolo y apagándolo, se recomienda al usuario que intente corregir la interferencia mediante una o más de las siguientes medidas: • Reorientar o reubicar la antena receptora. • Aumentar la separación entre el equipo y el receptor. • Conectar el equipo a un tomacorriente en un circuito diferente de aquel al que el receptor está conectado. • Consulte al distribuidor o a un técnico de radio / TV con experiencia para obtener ayuda. Nota de Industry Canada Este aparato digital Clase B cumple con la norma canadiense ICES-001. TABLE DES MATIÈRES 1 INTRODUCTION ..............................................................1 1.1 Mesures de sécurité ............................................................................. 1 2 INSTALLATION................................................................3 2.1 Contenu de l’emballage ....................................................................... 3 2.2 Remplacement ou installation du rotor ................................................ 4 2.3 Mise sous tension ................................................................................ 5 2.4 Préparation à la centrifugation ............................................................. 5 2.5 Centrifugation (en fonctionnement) ...................................................... 5 3 PRÉCAUTIONS PARTICULIÈRES ..................................6 3.1 Alimentation ......................................................................................... 6 3.2 Utilisation et maintenance .................................................................... 6 4 SPÉCIFICATIONS ............................................................7 5 CONFORMITÉ..................................................................8 FrontierTM FC5306 1 FR-1 INTRODUCTION Merci d'avoir choisi ce produit OHAUS. Veuillez lire le manuel intégralement avant d'utiliser la centrifugeuse TM Frontier FC5306 afin d'éviter un fonctionnement incorrect. Un grand nombre de fonctions utiles de la centrifugeuse de la série TM Frontier 5000 sont à découvrir. Veuillez trouver quelques-unes parmi les TM nombreuses caractéristiques actuelles de la série Frontier 5000:  Petit Boitier, beau désigne, faible encombrement et une seule     1.1 touche de fonctionnement L'installation (et /ou) le remplacement de rotor est facile , rapide, sans avoir besoin d'aucun outil Le rotor s'arrêtera immédiatement si le couvercle est soulevé pendant le fonctionnement L'adaptateur d'alimentation fonctionne dans la plage de 100 à 240 V CA Produit élaboré au bruit limité et à longue durée d'utilisation Mesures de sécurité PRÉCAUTIONS : Lire tous les avertissements de sécurité avant de procéder à l’installation, aux connexions ou à la maintenance de cet équipement. Si ces avertissements ne sont pas respectés, des accidents et/ou des dommages matériels peuvent en résulter. Conserver toutes les instructions pour référence ultérieure. Mesures de protection à respecter  Ne jamais travailler dans un environnement pouvant présenter des dangers d'explosion ! Le boîtier de l'instrument n'est pas étanche aux gaz (danger d'explosion due à la formation d'étincelles, corrosion provoquée par l'introduction de gaz).  Avec des produits chimiques et des solvants, se conformer aux instructions du fabricant des produits chimiques et aux règles générales de sécurité du laboratoire. FR-2 FrontierTM FC5306 Mesures à respecter pour la sécurité du fonctionnement  Ne pas s'appuyer contre la centrifugeuse.  Ne pas rester dans la zone de sécurité autour de la centrifugeuse plus que nécessaire pour des raisons d'exploitation.  Ne laisser aucun produit potentiellement dangereux dans la zone de sécurité autour de la centrifugeuse.  Ne pas dévisser les deux moitiés de l'enceinte.  Sécher immédiatement tout déversement de liquide. L'instrument n'est pas étanche.  Vérifier que la plage de tension d'entrée de l'adaptateur CA et le type de prise sont compatibles avec l'alimentation alternative locale devant être utilisée.  Ne pas positionner l'équipement pour qu'il soit difficile de déconnecter l'adaptateur CA de la prise d'alimentation.  S'assurer que le cordon d'alimentation n’est pas un obstacle potentiel ou ne présente aucun danger de chute.  L'équipement ne doit être utilisé qu'en intérieur. Utiliser l'équipement seulement dans des lieux secs.  Utiliser uniquement les accessoires et les périphériques approuvés.  Faire fonctionner l'équipement uniquement dans les conditions ambiantes spécifiées dans ces instructions.  Déconnecter l'équipement de l'alimentation électrique pour le nettoyer.  Ne pas utiliser l'équipement dans des environnements dangereux ou instables.  La maintenance doit être exclusivement exécutée par un personnel autorisé. Exclure les influences environnementales suivantes :  Vibrations puissantes  Exposition au Rayon de soleil  Humidité atmosphérique supérieure à 80 %  Présence de gaz corrosifs  Températures supérieures à 40 °C et inférieures à 5 °C  Champs électriques ou magnétiques puissants FrontierTM FC5306 2 FR-3 INSTALLATION Déballer soigneusement la boîte. 2.1 Contenu de l’emballage Le modèle Frontier Frontier TM TM FC5306 contient les articles suivants : FC5306 Photo Unités Machine principale TM Frontier FC5306 1 ensemble Rotor angulaire ( 8 x 1.5/2ml) 1 ensemble Rotor rectangulaire de 4 x 0.2 ml pour un ensemble de tubes PCR 8 1 ensemble Adaptateur dédié d'alimentation continue 24 V 1 ensemble Huit adaptateurs 0.5 ml et huit adaptateurs 0.2 ml 8 x 2 unités Manuel 1 ensemble FrontierTM FC5306 FR-4 2.2 1. Remplacement ou installation du rotor Lorsque la machine est livrée depuis l'usine, le manchon de l'axe de saisie et le rotor angulaire sont fermement fixés sur l'axe du moteur d'entraînemen. 2. S'il s'avère nécessaire de démonter un rotor installé, l'utilisateur doit maintenir le bord extérieur du rotor et pousser sur la protubérance du collier de serrage vers son centre avec les deux pouces, et ensuite soulever le rotor. 3. Pour réinstaller le rotor, aligner le trou ovale de rotor par-dessus le collier de serrage puis Appuyez doucement sur le rotor jusqu'à ce que le collier s'enclenche. 角转头 Rotor angulaire Collier de serrage Manchon de l'arbre moteur 电机轴套 FrontierTM FC5306 2.3 1. 2. 2.4 1. 2. 2.5 1. 2. 3. 4. FR-5 Mise sous tension Connecter la sortie de l'adaptateur sur la prise à l'arrière de la machine et brancher l'unité à une source d'alimentation. Mettre sous tension avec le bouton MARCHE ARRET à l'arrière de la centrifugeuse. Préparation à la centrifugation Ouvrir le couvercle transparent (voir la photo ci-dessous) et introduire la charge correctement de manière équilibrée (reportezvous à la section 3.2, Utilisation et maintenance). Refermer le couvercle et appuyer sur le contacteur d'alimentation MARCHE à l'arrière de cette machine (remarque : le rotor ne fonctionnera pas tant que le couvercle est ouvert). Centrifugation (en fonctionnement) Le couvercle étant fermé et la machine sous tension, le rotor commencera automatiquement sa rotation à vitesse croissante jusqu'à sa valeur nominale, puis se stabilisera. Ne pas ouvrir le couvercle de la centrifugeuse tant que la centrifugation n'est pas finie et attendre jusqu'à l'arrêt complet du rotor. Il n'est pas nécessaire de mettre l'appareil hors tension si une autre centrifugation s'avère nécessaire. Mettre hors tension et débrancher la machine lorsque tout le travail de centrifugation est terminé ou lorsqu'elle n'est pas en utilisation. Appuyer sur le couvercle pour l'ouvrir FrontierTM FC5306 FR-6 3 3.1 1. 2. 3.2 1. 2. 3. 4. 5. 6. PRÉCAUTIONS PARTICULIÈRES Alimentation L'alimentation de cette machine doit s'effectuer avec l'adaptateur l'accompagnant. La prise CA de l'adaptateur ne doit pas être connectée sur une alimentation triphasée. Ne Brancher, ni Débrancher la prise d'alimentation de l'adaptateur avec une main mouillée. Ne pas tirer sur le câble pour le débrancher de la prise. Utilisation et maintenance L'ajout d'un échantillon n'est autorisé que jusqu'à environ 70 % de la capacité nominale.Vérifier visuellement l'équilibrage lors de l'ajout d'un échantillon. Ne pas segmenter aucun échantillon biologique infectieux ou toxique. Ne pas segmenter aucun échantillon de produits chimiques explosifs ou inflammables. Les tubes à essai avec un échantillon à centrifuger doivent être positionnés symétriquement dans les orifices, afin de conserver l'équilibrage du rotor Lorsque le rotor est installé ou remplacé, toute force latérale excessive doit être évitée afin d'empêcher sa déformation ou son cintrage. Si le manchon de l'axe de saisie autour de l'axe du moteur d'entraînement est légèrement desserré, il est nécessaire de resserrer la vis de pression M3. Ne jamais utiliser un rotor brisé ou déformé. Au cours de la maintenance et du nettoyage quotidiens, la machine doit être essuyée à l'eau propre et avec un détergent neutre et être séchée par ventilation naturelle. Pendant l'utilisation, prendre toutes les précautions afin de conserver le positionnement horizontal et symétrique des tubes de centrifugation. AVERTISSEMENT : Déconnecter l'adaptateur CA de l'alimentation secteur et débrancher l'équipement avant le nettoyage. Ne laisser aucun liquide pénétrer dans l'adaptateur CA ou dans l'équipement. FrontierTM FC5306 4 FR-7 SPÉCIFICATIONS Les données techniques sont valides dans les conditions ambiantes suivantes : Température de fonctionnement : 5 à 40 °C Humidité relative : humidité relative entre 20 et 85 %, sans condensation Altitude au-dessus du niveau de la mer : Jusqu'à 2000 m Degré de pollution : 2 Catégorie d'installation : II Variations secteur de ±10 % de la tension nominale Modèle Mini Centrifugeuse Frontier :FC5306 Rotor angulaire Rotor réceptacle des tubes à essai Rotor Spécifications de la vitesse de rotation Force centrifuge Sens de la rotation Capacité maximum Adaptateur 6000 rpm ± 5% 2000 g ± 10% contre les aiguille d'une montre Rotor rectangulaire pour 8 PCR de 4 x 0.2 ml 8 x 1.5 ml / 2 ml 32 x 0.2 ml tube de centrifugation à usage unique 0.2 ml / 0.5 ml - Durée de l'accélération ≤2s Durée de la décélération ≤3s Moteur et mode d'entraînement Alimentation Dimensions Poids net Dimensions d'expédition Poids à l'expédition Entraînement direct, avec faible bruit, à cause des roulements Entrée : 100-240 V ~ 50/60Hz 1.5 A .6 A Sortie : 24 V  150 L x 168 P x 128 H mm 5.9 L x 6.6 P x 5 H po 1.05 Kg (comprenant le rotor angulaire hors charge) 2.316 lbs. 8.5 × 12.0 × 7.6 in. / 216 × 305 × 194 mm 4.079 lb / 1.85 kg FrontierTM FC5306 FR-8 5 CONFORMITÉ La conformité aux normes suivantes est indiquée par le marquage correspondant sur le produit. Marquage Standard Ce produit est conforme à la directive EMC 2004/108/EC et à la directive de basse tension 2006/95/EC. L'ensemble de la Déclaration de conformité est disponible en ligne sur http://europe.ohaus.com/europe/en/home/support/compliance/c e-declaration-of-conformity.aspx Dechets Conformément à la directive européenne 2002/96/CE sur les déchets d’équipements électriques et électroniques (WEEE), cet appareil ne peut pas être éliminé avec les déchets domestiques.. Ceci est également valable pour les pays en dehors de l’UE, selon leurs spécifications particulières. S'il vous plaît de se débarrasser de ce produit conformément à la réglementation locale, au point de collecte spécifié pour les équipements électriques et électroniques. En cas de questions, veuillez contacter l’autorité responsable ou le distributeur auprès duquel cet appareil a été acheté. Si cet appareil change de propriétaire (pour des raisons personnelles ou professionnelles), cette consigne doit lui être communiquée. Pour des instructions de mise au rebut en Europe, voir http://europe.ohaus.com/europe/en/home/support/weee.aspx. Nous vous remercions de votre contribution à la protection de l’environnement. FrontierTM FC5306 FR-9 Remarque FCC Cet équipement a été testé et déclaré conforme aux limites des appareils numériques de classe B, selon la Section 15 des règles du FCC (Commission fédérale des communications). Ces limites ont pour objectif de fournir une protection raisonnable contre des interférences dangereuses lorsque l'équipement est utilisé dans un environnement commercial. Cet équipement génère, utilise et peut rayonner des hautes fréquences et s'il n'est pas installé et utilisé conformément au guide de l'utilisateur, il peut générer des interférences préjudiciables aux communications radio. Cependant, il n'existe aucune garantie que des interférences ne se produiront pas dans une installation particulière. Si cet équipement provoque des interférences nuisibles à la réception radio ou télévision, ce qui peut être déterminé en mettant l'équipement hors et sous tension, l'utilisateur est encouragé à essayer de corriger l'interférence par une ou plusieurs des mesures suivantes: • Réorienter ou déplacer l'antenne de réception. • Augmenter la distance entre l'équipement et le récepteur. • Connecter l'équipement à une sortie sur un circuit différent de celui lequel le récepteur est connecté. • Consulter le revendeur ou un technicien radio / TV expérimenté. Remarque d’Industrie Canada Cet appareil numérique de classe B est conforme à la norme canadienne ICES-001. INHALTSVERZEICHNIS 1 EINLEITUNG ....................................................................................... 1 1.1 Sicherheitsmaßnahmen ....................................................................... 1 2 INSTALLATION ................................................................................... 3 2.1 Verpackungsinhalt ................................................................................ 3 2.2 Wechsel oder Installation des Rotors .................................................. 4 2.3 Einschalten ........................................................................................... 5 2.4 Vorbereitung für die Zentrifugierung .................................................... 5 2.5 Zentrifugierung (Betrieb) ...................................................................... 5 3 WICHTIGER HINWEIS ........................................................................ 6 3.1 Stromversorgung .................................................................................. 6 3.2 Verwendung und Wartung ................................................................... 6 4 TECHNISCHE DATEN ........................................................................ 7 5 EINHALTUNG ...................................................................................... 8 FrontierTM FC5306 1 DE-1 EINLEITUNG Vielen Dank, dass Sie sich für dieses OHAUS-Produkt entschieden haben. TM Bitte lesen Sie das Handbuch vollständig durch, bevor Sie die Frontier Zentrifuge FC5306 verwenden, um Fehler bei der Bedienung zu vermeiden. TM Die Zentrifugen der Frontier 5000-Serie sind mit vielen nützlichen Funktionen ausgestattet. Dazu gehören beispielsweise:  Kleines und durchdachtes Gehäuse, Eintastenbedienung  Schnelle und einfache Installation sowie schneller Rotorwechsel    1.1 ohne Werkzeug Der Rotor stoppt sofort, wenn der Deckel während des Betriebs geöffnet wird. Das Tischnetzteil ist für 100-240 V~ geeignet. Ausgereiftes, geräuscharmes Produkt mit langen Wartungsintervallen Sicherheitsmaßnahmen ACHTUNG: Lesen Sie zuerst alle Sicherheitshinweise, bevor Sie das Gerät installieren, anschließen oder reparieren. Die Nichtbeachtung der Warnhinweise kann zu Verletzungen und/oder Beschädigungen führen. Bewahren Sie alle Anweisungen zum späteren Nachschlagen auf. Maßnahmen zu Ihrem Schutz  Arbeiten Sie nie in einer explosiven Umgebung! Das Gehäuse des Geräts ist nicht gasdicht. (Es besteht Explosionsgefahr aufgrund von Funkenbildung und Korrosionsgefahr durch das Eindringen von Gasen)  Beachten Sie bei Verwendung von Chemikalien und Lösungsmitteln die Hinweise des Herstellers sowie die allgemeinen Laborsicherheitsvorschriften. DE-2 FrontierTM FC5306 Maßnahmen für die Betriebssicherheit  Lehnen Sie sich nicht an die Zentrifuge.  Bleiben Sie nicht länger im Sicherheitsbereich der Zentrifuge als für die Bedienung notwendig.  Lagern Sie keine potenziell gefährlichen Materialien innerhalb des Sicherheitsbereichs der Zentrifuge.  Schrauben Sie die beiden Hälften des Gehäuses nicht ab.  Wischen Sie verschüttete Flüssigkeiten sofort auf! Das Gerät ist nicht wasserdicht.  Prüfen Sie, ob die Eingangsspannung und der Netzstecker des Netzteils zu der Netzspannung vor Ort passen.  Stellen Sie das Gerät so auf, dass Sie das Netzteil problemlos von der Steckdose trennen können.  Das Netzkabel darf keine Stolpergefahr und kein Hindernis darstellen.  Das Gerät ist nur für die Verwendung in Räumen vorgesehen. Verwenden Sie das Gerät nur an trockenen Standorten.  Verwenden Sie nur zugelassene Zubehörteile und Peripheriegeräte.  Verwenden Sie das Gerät nur unter den in der Anleitung angegebenen Umgebungsbedingungen.  Trennen Sie die Waage vor dem Reinigen von der Stromversorgung.  Verwenden Sie die Waage nicht in gefährlichen oder instabilen Umgebungen.  Die Waage darf nur von autorisierten Mitarbeitern gewartet und repariert werden. Vermeiden Sie die folgenden Umwelteinflüsse:  Starke Schwingungen  Direktes Sonnenlicht  Luftfeuchtigkeit über 80 %  Korrosive Gase  Temperaturen unter 5 °C und über 40 °C  Starke elektrische oder magnetische Felder FrontierTM FC5306 2 DE-3 INSTALLATION Nehmen Sie die Teile sorgfältig aus der Verpackung. 2.1 Verpackungsinhalt Das Modell Frontier Frontier TM TM FC5306 umfasst folgende Teile: FC5306 Foto Einheiten TM Frontier FC5306 Hauptgerät 1 Stück Winkelrotor (8 x 1,5 / 2 ml) 1 Stück 4 x 0,2ml rechteckiger Rotor für PCR 8Röhrchenreihe 1 Stück Spezielles 24 V-Netzteil 1 Stück Acht 0,5 ml-Adapter und acht 0,2 ml-Adapter 8 x 2 Einheiten Handbuch 1 Stück FrontierTM FC5306 DE-4 2.2 1. Wechsel oder Installation des Rotors Bei Lieferung vom Werk sind die Klemmhülse der Welle und der Winkelrotor fest auf der Welle des Antriebsmotors montiert. 2. Um einen installierten Rotor auszubauen, müssen Sie die Außenkante des Rotors festhalten, die hervorstehende Klemme mit beiden Daumen in die Mitte drücken und dann den Rotor nach oben heben. 3. Um einen Rotor einzusetzen, müssen Sie die ovale Bohrung des Rotors auf die Klemme ausrichten und dann den Rotor vorsichtig nach unten drücken, bis die Klemme einrastet. Winkelrotor 角转头 Klemme Motorwellenhülse 电机轴套 FrontierTM FC5306 2.3 1. 2. 2.4 1. 2. 2.5 1. 2. 3. 4. DE-5 Einschalten Sie können das Gerät mit dem Netzschalter auf der Rückseite ausschalten. Verbinden Sie das Ausgangsende des Netzteils mit dem Netzeingang auf der Rückseite der Zentrifuge und schließen Sie dann das Gerät an eine Stromquelle an. Vorbereitung für die Zentrifugierung Öffnen Sie den durchsichtigen Deckel (siehe folgendes Foto) und setzen Sie die Proben symmetrisch ein (siehe Abschnitt “3.2 Verwendung und Wartung“) Schließen Sie den Deckel und schalten Sie den Netzschalter auf der Rückseite des Geräts ein (Hinweis: Der Rotor dreht sich nicht, solange der Deckel geöffnet ist). Zentrifugierung (Betrieb) Sobald der Deckel geschlossen und das Gerät eingeschaltet ist, beginnt der Rotor automatisch mit zunehmender Geschwindigkeit zu drehen. Wenn die Nenngeschwindigkeit erreicht ist, dreht sich der Rotor mit dieser Geschwindigkeit weiter. Öffnen Sie den Deckel, wenn die erforderliche Zentrifugierung abgeschlossen ist; der Rotor verlangsamt sich sofort, bis er vollständig steht. Sie müssen das Gerät nicht erst ausschalten, wenn eine weitere Zentrifugierung notwendig ist. Schalten Sie das Gerät aus und ziehen Sie den Netzstecker, wenn die Zentrifugierungsarbeiten abgeschlossen sind oder das Gerät nicht verwendet wird. Nach unten drücken, um den Deckel zu öffnen. FrontierTM FC5306 DE-6 3 3.1 1. 2. 3.2 1. 2. 3. 4. 5. 6. WICHTIGER HINWEIS Stromversorgung Die Stromversorgung des Geräts erfolgt über das Netzteil, das im Lieferumfang des Geräts enthalten ist. Der Stecker des Netzteils darf nicht mit einer dreiphasigen Steckdose verbunden werden. Verbinden oder trennen Sie den Stecker des Netzteils nicht mit nassen Händen. Ziehen Sie nicht am Kabel, um den Stecker herauszuziehen. Verwendung und Wartung Bereiten Sie die Mikro-Bürette der Standardspezifikation mit dem Klappdeckel auf die Zentrifugierung vor. Sie dürfen die Bürette nur bis ca. 70 % der Nennkapazität mit Proben füllen. Prüfen Sie die symmetrische Anordnung beim Einsetzen der Proben visuell. Achten Sie darauf, dass kein giftiges oder infektiöses Probenmaterial austritt. Achten Sie darauf, dass kein entzündliches oder explosives chemisches Probenmaterial austritt. Die Zentrifugenröhrchen mit dem Probenmaterial müssen symmetrisch in die Röhrchenaussparungen des Zentrifugenrotors platziert werden, damit der Rotor ausgewuchtet bleibt. Vermeiden Sie beim Einsetzen oder Ersetzen des Rotors eine zu starke Seitenkraft, damit der Rotor sich nicht verformt oder verbiegt. Wenn die Klemmhülse um die Motorwelle etwas locker ist, muss die M3-Stellschraube etwas nachgezogen werden. Verwenden Sie keinen gebrochenen oder verformten Rotor. Wischen Sie das Gerät bei der täglichen Reinigung und Wartung mit sauberem Wasser oder einem neutralen Reinigungsmittel ab und lassen Sie es an der Luft trocknen. Achten Sie bei der Verwendung besonders auf die symmetrische und horizontale Platzierung der Zentrifugenröhrchen. VORSICHT: Trennen Sie das Netzteil vor der Reinigung von der Steckdose und vom Gerät. Es dürfen keine Flüssigkeiten in das Netzteil oder das Gerät eindringen. FrontierTM FC5306 4 DE-7 TECHNISCHE DATEN Die technischen Daten gelten unter folgenden Umgebungsbedingungen: Betriebstemperatur: 5 °C bis 40 °C Relative Luftfeuchtigkeit: 20 % bis 85 % relative Luftfeuchtigkeit, nicht kondensierend Höhe über Meeresspiegel: bis 2000 m Verschmutzungsgrad: 2 Installationskategorie: II Netzspannungsschwankungen: bis ± 10 % der Nennspannung Modell Rotor Max. Rotationsgeschwindigkeit Frontier-Minizentrifuge:FC5306 Winkelrotor Röhrchenreihenrotor 6000 U/min ± 5% Zentrifugalkraft 2000 g ± 10% Rotationsrichtung Gegen den Uhrzeigersinn Max. Kapazität 8 x 1,5 ml / 2 ml 4 x 0,2 ml-PCR 8Röhrchenreihe oder 32 x 0,2 ml einzelne Zentrifugenröhrchen Netzteil 0,2 ml / 0,5 ml - Beschleunigung ≤2s Abbremsung ≤3s Antriebsmodus und Motor Direktantrieb, geräuscharm dank Rollenlager Stromversorgung Eingang: 100–240 V~, 50/60 Hz, 1,5 A Ausgang: 24 V–  .6 A Abmessungen (B × T × H) 5,9 × 6,6 × 5,0 in. / 150 × 168 × 128 mm Nettogewicht 2,316 lb / 1,05 kg (mit Freilast-Winkelrotor) Versandabmessungen (B × T × H) 8,5 × 12,0 × 7,6 in. / 216 × 305 × 194 mm Versandgewicht 4,079 lb / 1,85 kg Frontier TM FC5306 DE-8 5 EINHALTUNG Die Einhaltung folgender Normen ist durch eine entsprechende Kennzeichnung auf dem Produkt vermerkt. Kennzeichnung Norm Dieses Produkt entspricht der EMV-Richtlinie 2004/108/EG und der Niederspannungsrichtlinie 2006/95/EG. Die vollständige Konformitätserklärung finden Sie online unter http://europe.ohaus.com/europe/en/home/support/compliance/c e-declaration-of-conformity.aspx Entsorgung Entsprechend der Europäischen Richtlinie 2002/96/EG über Elektro- und Elektronik-Altgeräte (WEEE) darf dieses Gerät nicht im Hausmüll entsorgt werden. Dies gilt auch für Länder außerhalb der EU entsprechend den jeweiligen Vorschriften. Entsorgen Sie dieses Produkt entsprechend den lokalen Vorschriften an der Sammelstelle für Elektro- und Elektronikgeräte. Bei Fragen wenden Sie sich bitte an die entsprechende Behörde oder an den Händler, von dem Sie das Gerät erworben haben. Wenn Sie das Gerät an andere weitergeben (zur privaten oder beruflichen Nutzung), muss der Inhalt dieser Vorschriften ebenfalls weitergeben werden. Entsorgungshinweise für Europa finden Sie unter http://europe.ohaus.com/europe/en/home/support/weee.aspx. Vielen Dank für Ihren Beitrag zum Umweltschutz. FrontierTM FC5306 DE-9 FCC-Hinweis Jedes Gerät wurde geprüft und entspricht den Grenzwerten für Digitalgeräte der Klasse B gemäß Abschnitt 15 der FCC-Vorschriften. Diese Grenzwerte sollen in Wohnumgebungen einen angemessenen Schutz gegen Funkstörungen gewährleisten. Dieses Gerät erzeugt und verwendet HF-Energie und strahlt diese u. U. auch ab; wird es nicht entsprechend der Anweisung installiert und verwendet, können Funkstörungen auftreten. Es gibt jedoch keine Garantie, dass bei einer bestimmten Installation keine Störung auftritt. Wenn dieses Gerät zu Funkstörungen beim Radio- oder Fernsehempfang führt, was sich durch Ein- und Ausschalten des Geräts überprüfen lässt, sollte der Benutzer die Störung durch eine oder mehrere der folgenden Maßnahmen beheben: • Neuausrichtung oder Umsetzung der Empfangsantenne • Größerer Abstand zwischen Gerät und Empfänger • Anschluss des Geräts an eine Steckdose in einem anderen Stromkreis als der Empfänger • Beratung durch den Händler oder einen erfahrenen Radio/Fernsehtechniker Industry Canada-Hinweis Das Digitalgerät der Klasse B entspricht den kanadischen Vorschriften ICES-001. INDICE 1 INTRODUZIONE ..........................................................1 1.1 Misure di sicurezza .............................................................................. 1 2 INSTALLAZIONE ........................................................3 2.1 Contenuto della confezione ................................................................. 3 2.2 Sostituzione o installazione del rotore ................................................. 4 2.3 Accensione ........................................................................................... 5 2.4 Preparazione per la centrifuga ............................................................. 5 2.5 Centrifuga (In esecuzione) ................................................................... 5 3 ATTENZIONE SPECIALE .........................................6 3.1 Alimentazione ....................................................................................... 6 3.2 Uso e manutenzione ............................................................................ 6 4 SPECIFICHE TECNICHE ..........................................7 5 CONFORMITÀ .............................................................8 FrontierTM FC5306 FrontierTM FC5306 1 IT-1 EN-1 INTRODUZIONE Grazie per aver scelto questo prodotto OHAUS. Prima di usare Frontier™ FC5306 leggere il manuale interamente per evitare operazioni non corrette. TM La centrifuga della serie Frontier 5000 include molte funzioni utili. Alcune TM tra queste funzioni per Frontier FC5306 sono:  Design piccolo e molto bello, operatività con un solo tasto  Installazione conveniente e rapida e sostituzione del rotore senza    1.1 attrezzi Il movimento del rotore si interrompe immediatamente se durante il funzionamento viene aperto il coperchio L'adattatore di alimentazione del desktop è applicabile a alimentazioni da 100-240 V CA Prodotto elaborato poco rumoroso e con vita utile lunga Misure di sicurezza ATTENZIONE: leggere tutte le avvertenze di sicurezza prima di procedere all'installazione, alla realizzazione dei collegamenti o alla manutenzione dell'apparecchiatura. Il mancato rispetto delle avvertenze suddette può provocare lesioni personali e/o danni alla proprietà. Conservare tutte le istruzioni per poterle consultare in caso di necessità. Misure di protezione personale  Non lavorare mai in un ambiente soggetto a pericolo di esplosione! L'alloggiamento dello strumento non è a prova di gas. (pericolo di esplosione dovuto a formazione di scintille, corrosione causata da ingresso di gas)  Quando si utilizzano prodotti chimici e solventi, seguire le istruzioni del produttore e le regole di sicurezza presenti sull'etichetta generica. FrontierTM FC5306 IT-2 Misure per la sicurezza operativa  Non appoggiarsi sulla centrifuga.  Non rimanere all'interno dell'area di sicurezza più a lungo del necessario per motivi operativi.  Non depositare materiali potenzialmente pericolosi all'interno dell'area di sicurezza.  Non svitare le metà dell'alloggiamento.  Asciugare immediatamente eventuali liquidi versati! Lo strumento non è impermeabile.  Verificare che l’intervallo di tensione in ingresso dell'adattatore CA e il tipo di spina siano adatti all'alimentazione elettrica locale.  Non posizionare l'apparecchiatura in modo tale che risulti difficoltoso scollegare l'adattatore CA dalla presa elettrica.  Accertarsi che il cavo di alimentazione non rappresenti un potenziale ostacolo o non costituisca un intralcio al passaggio.  Quest'apparecchiatura è destinata esclusivamente all'uso interno. Utilizzare l'apparecchio solo in luoghi asciutti.  Utilizzare solo accessori e periferiche approvate.  Utilizzare l’apparecchiatura solo nelle condizioni ambientali specificate in queste istruzioni.  Prima di effettuare la pulizia dell’apparecchiatura, scollegare l'alimentazione elettrica.  Non utilizzare l’apparecchiatura in ambienti instabili o pericolosi.  La manutenzione deve essere eseguita soltanto da personale autorizzato. Escludere le seguenti influenze ambientali:       Vibrazioni potenti Luce solare diretta Umidità atmosferica superiore all'80% Presenza di gas corrosive Temperature inferiori a 5 °C e superiori a 40 °C Campi elettrici o magnetici potenti FrontierTM FC5306 IT-3 2 INSTALLAZIONE 2.1 Contenuto della confezione Disimballaggio con cura. Il modello Frontier Frontier TM TM FC5306 presenta i seguenti elementi: FC5306 Foto Unità Macchina principale TM Frontier FC5306 1 set Rotore angolare (8 x 1,5 / 2ml) 1 set Rotore rettangolare 4 x 0,2ml per banco a 8 provette in PCR 1 set Alimentatore dedicato CC 24 V 1 set Otto adattatori da 0,5 ml e otto adattatori da 0,2 ml Manuale 8 x 2 unità 1 set FrontierTM FC5306 IT-4 2.2 1. Sostituzione o installazione del rotore Quando la macchina viene consegnata dalla fabbrica, il manicotto dell'albero di blocco e il rotore angolare sono fermi in modo fisso sull'albero del motore a trazione. 2. Se è necessario smontare un rotore installato, l'utente deve mantenere il bordo esterno del rotore e spingere il morsetto che protrude verso il centro con i pollici, quindi sollevare in alto il rotore. 3. Per reinstallare il rotore, allineare il foro ovale del rotore sopra al morsetto, quindi premere delicatamente verso il basso il rotore fino a quando il morsetto non scatta in posizione. 角转头 Rotore angolare Morsetto Manicotto albero motore 电机轴套 FrontierTM FC5306 2.3 1. 2. 2.4 1. 2. 2.5 1. 2. 3. 4. IT-5 Accensione Con l'interruttore di alimentazione sul retro della macchina su OFF. Collegare l'estremità di uscita dell'adattatore nella presa sul retro della macchina e quindi collegare a una presa elettrica. Preparazione per la centrifuga Aprire il coperchio trasparente (vedere foto sotto) e aggiungere in modo corretto il carico e bilanciarlo (fare riferimento alla sezione “3.2 uso e manutenzione”). Chiudere il coperchio e premere l'interruttore di alimentazione sul retro della macchina in posizione ON (nota: il rotore non funziona con il coperchio aperto). Centrifuga (In esecuzione) Dopo aver chiuso il coperchio a acceso la macchina, il rotore inizia automaticamente a funzionare aumentando la velocità fino alla velocità nominale e quindi funziona in modo stabile. Aprire il coperchio quando sono soddisfatti i requisiti di centrifugazione, il rotore rallenta immediatamente fino all'arresto completo. Non è necessario spegnere se si deve fare un'altra centrifugazione. Spegnere e scollegare la macchina dalla presa elettrica al termine del lavoro di centrifugazione o se non è in uso. Premere verso il basso per aprire il coperchio FrontierTM FC5306 IT-6 3 3.1 1. 2. 3.2 1. 2. 3. 4. 5. 6. ATTENZIONE SPECIALE Alimentazione Questa macchina deve essere alimentata tramite l'adattatore fornito. La spina CA dell'adattatore non deve essere collegata a fonti elettriche trifase. Non inserire la spina o l'adattatore con le mani bagnate. Non tirare con forza il cavo per staccare la spina. Uso e manutenzione Preparare la micro buretta con le specifiche standard con il coperchio con cerniera per la centrifugazione. È possibile aggiungere campione fino a circa il 70% della capacità nominale al massimo. Controllare visivamente il bilanciamento quando si aggiunge il campione. Non staccare campioni biologici velenosi o infetti. Non staccare campioni chimici infiammabili o velenosi. Le provette per la centrifuga con campione devono essere posizionate in modo simmetrico nei fori per provetta per centrifuga del rotore per mantenere il rotore bilanciato. Quando viene installato o sostituito il rotore, evitare un'eccessiva forza laterale per impedire deformazioni o piegature. Se il manicotto dell'albero di blocco attorno all'albero del motore è un poco lento, è necessario serrare di nuovo la vite M3. Non utilizzare un rotore rotto o deformato. Durante la pulizia e la manutenzione giornaliere, la macchina deve essere pulita con acqua pulita o detergente neutro e asciugata tramite ventilazione naturale. Durante l'uso, fare attenzione a posizionare in modo simmetrico e orizzontale le provette per la centrifuga. AVVERTENZA: Prima di effettuare la pulizia dell’apparecchiatura, scollegare l'adattatore Ca dall'alimentazione elettrica CA. Non lasciare che liquidi penetrino nell'adattatore CA o nell'apparecchio. FrontierTM FC5306 4 IT-7 SPECIFICHE TECNICHE I dati tecnici sono validi per le seguenti condizioni ambientali: Temperatura di funzionamento: da 5℃ a 40 ℃ Umidità relativa: 20-85% di umidità relativa, senza condensa Altitudine s.l.m.: Fino a 2.000 m Grado di inquinamento: 2 Categoria d'installazione: II Fluttuazioni di tensione di alimentazione minima: fino a ±10% della tensione nominale Modello Rotore Centrifuga Frontier Mini:FC5306 Rotore con banco Rotore angolare provette Velocità di rotazione 6000 giri/min ± 5% Forza centrifuga 2000 g ± 10% Direzione di rotazione Senso antiorario Capacità Max 8 x 1.5 ml / 2 ml banco per 8 provette 4 x 0,2 ml PCR o per provetta centrifuga singola 32 x 0,2 ml Adattatore 0,2 ml/0,5 ml - Tempo di accelerazione ≤2s Tempo di decelerazione Modalità trasmissione e motore ≤3s Trasmissione diretta poco rumore dovuto alle sfere rotanti Alimentazione Ingresso: 100-240 V ~ 50/60Hz 1.5A Uscita: 24V  .6 A Dimensioni (L x A x P) 150 × 168 × 128 mm / 5.9 × 6,6 × 5,0" Peso netto 1,05 kg / 2.316 lb (incluso rotore angolo di caricamento libero) Dimensioni dell’imballo (L × A × P) 216 × 305 × 194 mm / 8,5 × 12.0 × 7,6" Peso di spedizione 1,85 kg / 4.,79 lb FrontierTM FC5306 IT-8 5 CONFORMITÀ La conformità agli standard seguenti viene indicata dalla presenza del marchio corrispondente sul prodotto. Marchio Standard Questo prodotto è conforme alla direttiva EMC 2004/108/CEE e alla Direttiva Bassa Tensione 2006/95/CE. La Dichiarazione di Conformità è disponibile online all'indirizzo http://europe.ohaus.com/europe/en/home/support/compliance/c e-declaration-of-conformity.aspx Smaltimento In conformità alla Direttiva Europea 2002/96/CE sullo smaltimento delle apparecchiature elettriche ed elettroniche (RAEE), l’apparecchiatura non deve essere smaltita assieme ad altri rifiuti domestici. Queste disposizioni sono valide anche nei paesi esterni all'UE, in base ai requisiti delle varie legislazioni. Smaltire questo prodotto secondo le normative locali, presso il punto di raccolta specificato per le apparecchiature elettriche ed elettroniche. In caso di dubbi, rivolgersi all'ente responsabile o al distributore da cui è stato acquistato questo dispositivo. Qualora l’apparecchiatura sia trasferita a terzi (per uso privato o professionale), il contenuto della presente normativa deve ritenersi valido anche per terzi. Per le istruzioni relative allo smaltimento in Europa, fare riferimento a http://europe.ohaus.com/europe/en/home/support/weee.aspx. Grazie per il contributo alla salvaguardia dell'ambiente.conformance. FrontierTM FC5306 IT-9 Nota per FCC Questa apparecchiatura è stata collaudata e ritenuta conforme ai limiti dei dispositivi digitali di Classe B, in conformità con la Parte 15 delle norme FCC. Tali limiti sono previsti per fornire un’adeguata protezione contro interferenze dannose durante il funzionamento dell'apparecchiatura in un'installazione residenziale. L’apparecchiatura genera, utilizza e può irradiare energia nella banda delle radiofrequenze, per cui, se non installata e utilizzata in conformità alle istruzioni, potrebbe causare interferenze dannose con le comunicazioni radio. Tuttavia, non vi è garanzia che l'interferenza non si verifichi in una particolare installazione. Se tale apparecchiatura causa realmente interferenza pericolosa alla ricezione di radio e televisione, determinabile mediante lo spegnimento e la riaccensione dell'apparecchiatura, l'utente è incoraggiato a provare a correggere l'interferenza adottando una delle seguenti misure: • Riorientare o mettere in un altro posto l'antenna di ricezione. • Aumentare la distanza di separazione tra l'apparecchio e il ricevitore. • Collegare l'apparecchio a una presa su un circuito diverso da quello cui è collegato il ricevitore. • Consultare il rivenditore o un tecnico radio/TV esperto per aiuto. Nota per il Canada Questa apparecchiatura digitale di classe B è conforme alla normativa canadese ICES-001. Ohaus Corporation 7 Campus Drive Suite 310 Parsippany, NJ 07054 USA With offices worldwide / Con oficinas en todo el mundo / Avec des bureaux dans le monde entier / Mit weltweiten Niederlassungen / Con uffici in tutto il mondo http://www.ohaus.com *30129575* P/N 30129575 G © 2020 OHAUS Corporation, all rights reserved / todos los derechos reservados / tous droits réservés / Alle Rechte vorbehalten / tutti i diritti riservati Printed in China / Impreso en China / Imprimé en Chine / Gedruckt in China / Stampato in Cina
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52

Ohaus FC5306 Manual de usuario

Tipo
Manual de usuario
Este manual también es adecuado para