Ohaus ST350 Manual de usuario

Tipo
Manual de usuario
General instruction for pH electrodes
Instrucciones generales para electrodos de pH
Instructions gérales pour
les électrodes de pH
Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054 USA
Tel: (973) 377-9000
Fax: (973) 944-7177
With offices worldwide
Con oficinas en todo el mundo
Avec des bureaux dans le monde entier
www.ohaus.com; [email protected]
*30131311*
P/N 30131311 B © 2015
Ohaus Corporation, all rights reserved/ todos
los derechos reservados/ tous droits réservés
EN-2 ST350 electrodes
ST350 electrodes EN-1
Model
Item
NO
Description
ST350
30129354
3-in-1 pH electrode
Specifications:
Connection: BNC, Cinch
Cable length: 1.0m
Shaft length: 120mm
Shaft diameter: 12mm
pH range: 0-14pH
Zero potential: 7.00±0.25pH
Slope: 97%
Impedance: 100M
Shaft material: Glass
Reference Electrolyte:
3M KCl Solution
Temperature range: 0-100
Temperature probe: NTC 30kΩ
Reference Diaphragm: Sleeve
OHAUS offer different pH electrode for
different application. Make sure you have
chosen the best fit pH electrode for your
samples; and do calibration first for the
electrode.
3-in-1 electrodes have the temperature probe
in itthe diameter of the electrode should be
smaller than the tube.
EN-2 ST350 electrodes
Before using electrode in calibration or
measurement, we should note:
1. When unpacking, check the electrode for
any mechanical damage to the membrane,
diaphragm or body.
2. Screw off watering cap first, then the lid,
rinse electrode with pure water. Wipe
electrode with a tissue. Do not rub the glass
membrane, or the response time might
increase.
3. Eliminate any air bubbles inside the
membrane. Make the electrode inside
pressure equilibrium with the atmosphere by
opening the electrode refilling aperture when
measurement.
FAQs:
1. Whats the meaning of electrode
condition icon?
After calibration, the pH meter will judge the
electrode slope and offset, then display the
judgment in the screen:
slope > 95%
Electrode condition is very good
slope : 90%-95%
Electrode condition is acceptable
slope : 85%-90%
Electrode condition is not good, need re-
calibration or replace
ST350 electrodes EN-3
2. How do you properly store the pH
electrode after every use?
A: Rinse the electrode with pure water, put
the electrode into the watering cap. If the
solution in the cap is polluted or dry, clean it
and fill it with 3M KCl solution.
3.Whats the life of the pH electrodes?
A: Measured samples, storage conditions
and maintenance methods all could influence
the life of pH electrodes; normally a pH
electrode expires after 12 months. Some
sample, such as strong acid or alkali,
corrosive liquid will shorten the pH electrode
life. pH electrode is consumable, is generally
provided warranty.
4.why is the pH value not stable when
measuring pure water ?
A: generally, the common pH electrode
couldn't measure distilled water, rainwater or
pure water. Because they are the liquids,
which have low conductivity.
ST350 ES-1
Modelo
Artículo n°
Descripción
ST350
30129354
Electrodo de pH 3 en 1
Especificaciones:
Conexión: BNC, Cinch
Longitud del cable: 1.0 m
Longitud del eje: 120 mm
Diámetro del eje: 12 mm
Rango de pH: 0-14pH
Potencial cero: 7.00±0.25 pH
Pendiente: ≥ 97%
Impedancia: ≤ 100 MΩ
Material del eje: Vidrio
Electrólito de referencia:
Solución 3M KCl
Rango de temperatura: 0-100 ºC
Sonda de temperatura: NTC 30
Diafragma de referencia: Tubito
OHAUS ofrece diferentes electrodos de pH
para diferentes aplicaciones. Asegúrese de
elegir el mejor electrodo de pH para sus
muestras, y primero calibre el electrodo.
Los electrodos 3 en 1 tienen la sonda de
temperatura integrada
ES-2 ST350
Antes de usar el electrodo en calibracn o
medición, asegúrese de:
1. Al desempacar, revise el electrodo para
identificar cualquier daño mecánico en la
membrana, diafragma o cuerpo.
2. Destornille primero el tapón, luego la tapa,
y enjuague el electrodo con agua pura.
Limpie el electrodo con un pañuelo. No frote
la membrana de vidrio, ya que se puede
aumentar el tiempo de respuesta.
3. Elimine cualquier burbuja dentro de la
membrana. Equilibre la presión interna de la
membrana con la de la atmósfera abriendo el
orificio de relleno del electrodo al hacer una
medición.
Preguntas frecuentes:
1. ¿Qué significa el ícono de condición
del electrodo?
Después de la calibración, el medidor de pH
determinará la pendiente y desfase del
electrodo, y luego mostrará el resultado en la
pantalla:
Pendiente > 95%
La condición del electrodo es muy buena
Pendiente: 90%-95%
La condición del electrodo es aceptable
Pendiente: 85%-90%
La condición del electrodo no es buena;
necesita recalibrarse o reemplazarse.
ST350 ES-3
2. ¿mo se almacena adecuadamente el
electrodo de pH después de cada uso?
R: Enjuague el electrodo con agua pura,
coloque el electrodo en el tapón de
humedecimiento. Si la solución en el tapón
está contaminada o seca, límpielo y llénelo
con solución 3M KCl.
3. ¿Cuál es la vida útil de los electrodos
de pH?
R: Las muestras medidas, las condiciones de
almacenamiento y los métodos de
mantenimiento influyen en la vida útil de los
electrodos de pH; normalmente un electrodo
de pH caduca desps de 12 meses.
Algunas muestras, como los ácidos o álcalis
fuertes y los líquidos corrosivos acortan la
vida útil del electrodo. El electrodo de pH es
un consumible y generalmente se
proporciona garantía.
4. ¿Por qué el valor de pH no es estable
cuando se mide agua pura?
R: Generalmente, el electrodo de pH común
no puede medir agua destilada, agua de
lluvia o agua pura. Esto se debe a que son
quidos y poseen baja conductividad.
ST350 FR-1
Modèle
Article n°
Description
ST350
30129354
Électrode de pH 3 en 1
Spécifications :
Connexion : BNC, Cinch
Longueur du câble : 1 m
Longueur de l'axe : 120 mm
Diamètre de l'axe : 12 mm
Gamme de pH : 0-14pH
Potentiel zéro : 7,00 ±0,25 pH
Pente : ≥ 97 %
Impédance : 100 MΩ
Composition de l'axe : Verre
Électrolyte de rérence :
Solution 3M KCl
Plage de température : 0-100
Sonde de température : NTC 30
Diaphragme de rérence :
Manchon
OHAUS offre différentes électrodes de pH
pour différentes applications. S'assurer
d'avoir choisi l'électrode de pH la mieux
adaptée à vos échantillons ; procéder
d'abord à un étalonnage de l'électrode.
Les électrodes 3 en 1 contiennent une sonde
de température.
FR-2 ST350
Avant d'utiliser l'électrode pour un
étalonnage ou une mesure, ce qui suit doit
être noté :
1. Lors du déballage, vérifier l'électrode à la
recherche d'endommagement mécanique de
la membrane, du diaphragme ou du boîtier.
2. Dévisser d'abord le bouchon de mouillage,
puis le couvercle et rincer l'électrode à l'eau
pure. Essuyer l'électrode avec un papier-
mouchoir. Ne pas frotter la membrane de
verre, sinon le temps de réponse peut
augmenter.
3. Éliminer toutes les bulles d'air à l'intérieur
de la membrane. Procéder à l'équilibre de la
pression entre l'intérieur de l'électrode et
l'atmosphère en ouvrant l'ouverture de
remplissage de l'électrode lors de la mesure.
Questions fréquentes :
1. Quelle est la signification de l'icône
d'état de l'électrode ?
Après l'étalonnage, le pH-mètre jugera la
pente et le décalage de l'électrode avant
d'afficher son jugement sur l'écran :
pente > 95 %
L'état de l'électrode est très bon
pente : 90 %-95 %
L'état de l'électrode est acceptable
pente : 85 %-90 %
L'état de l'électrode n'est pas bon ou
cessite un nettoyage
ST350 FR-3
2. Comment pouvoir stocker correctement
l'électrode de pH après chaque utilisation ?
R : Rincer l'électrode à l'eau pure, ranger
l'électrode dans le bouchon de mouillage. Si
la solution dans le bouchon est polluée ou
che, le nettoyer et le remplir avec une
solution 3M KCl.
3. Quelle est la durée de vie d'une
électrode de température ?
R : Les échantillons mesurés, les conditions
de stockage et les méthodes de maintenance
peuvent tous influencer la durée de vie des
électrodes de pH ; normalement, une
électrode de pH devient périmée après 12
mois. Certains échantillons, comme des
liquides corrosifs, fortement acides ou
alcalins réduiront la durée de vie d'une
électrode de pH. Une électrode de pH
s'apparente à un consommable et est
ralement fournie avec une garantie.
4. Pourquoi la valeur du pH est-elle
instable lors de la mesure d'une eau pure ?
R: Géralement, une électrode classique de
pH ne peut pas mesurer l'eau distillée, l'eau
de pluie ou l'eau pure. Parce que ce sont des
liquides possédant une faible conductivité.

Transcripción de documentos

General instruction for pH electrodes Instrucciones generales para electrodos de pH Instructions générales pour les électrodes de pH Ohaus Corporation 7 Campus Drive Suite 310 Parsippany, NJ 07054 USA Tel: (973) 377-9000 Fax: (973) 944-7177 With offices worldwide Con oficinas en todo el mundo Avec des bureaux dans le monde entier www.ohaus.com; [email protected] *30131311* P/N 30131311 B © 2015 Ohaus Corporation, all rights reserved/ todos los derechos reservados/ tous droits réservés EN-2 ST350 electrodes ST350 electrodes Model ST350 Item NO 30129354 EN-1 Description 3-in-1 pH electrode Specifications: Connection: BNC, Cinch Cable length: 1.0m Shaft length: 120mm Shaft diameter: 12mm pH range: 0-14pH Zero potential: 7.00±0.25pH Slope: ≥ 97% Impedance: ≤ 100MΩ Shaft material: Glass Reference Electrolyte: 3M KCl Solution Temperature range: 0-100℃ Temperature probe: NTC 30kΩ Reference Diaphragm: Sleeve OHAUS offer different pH electrode for different application. Make sure you have chosen the best fit pH electrode for your samples; and do calibration first for the electrode. 3-in-1 electrodes have the temperature probe in itthe diameter of the electrode should be smaller than the tube. EN-2 ST350 electrodes Before using electrode in calibration or measurement, we should note: 1. When unpacking, check the electrode for any mechanical damage to the membrane, diaphragm or body. 2. Screw off watering cap first, then the lid, rinse electrode with pure water. Wipe electrode with a tissue. Do not rub the glass membrane, or the response time might increase. 3. Eliminate any air bubbles inside the membrane. Make the electrode inside pressure equilibrium with the atmosphere by opening the electrode refilling aperture when measurement. FAQs: 1. What’s the meaning of electrode condition icon? After calibration, the pH meter will judge the electrode slope and offset, then display the judgment in the screen: slope > 95% Electrode condition is very good slope : 90%-95% Electrode condition is acceptable slope : 85%-90% Electrode condition is not good, need recalibration or replace ST350 electrodes EN-3 2. How do you properly store the pH electrode after every use? A: Rinse the electrode with pure water, put the electrode into the watering cap. If the solution in the cap is polluted or dry, clean it and fill it with 3M KCl solution. 3.What’s the life of the pH electrodes? A: Measured samples, storage conditions and maintenance methods all could influence the life of pH electrodes; normally a pH electrode expires after 12 months. Some sample, such as strong acid or alkali, corrosive liquid will shorten the pH electrode life. pH electrode is consumable, is generally provided warranty. 4.why is the pH value not stable when measuring pure water ? A: generally, the common pH electrode couldn't measure distilled water, rainwater or pure water. Because they are the liquids, which have low conductivity. ST350 ES-1 Modelo ST350 Artículo n° 30129354 Descripción Electrodo de pH 3 en 1 Especificaciones: Conexión: BNC, Cinch Longitud del cable: 1.0 m Longitud del eje: 120 mm Diámetro del eje: 12 mm Rango de pH: 0-14pH Potencial cero: 7.00±0.25 pH Pendiente: ≥ 97% Impedancia: ≤ 100 MΩ Material del eje: Vidrio Electrólito de referencia: Solución 3M KCl Rango de temperatura: 0-100 ºC℃ Sonda de temperatura: NTC 30 kΩ Diafragma de referencia: Tubito OHAUS ofrece diferentes electrodos de pH para diferentes aplicaciones. Asegúrese de elegir el mejor electrodo de pH para sus muestras, y primero calibre el electrodo. Los electrodos 3 en 1 tienen la sonda de temperatura integrada ES-2 ST350 Antes de usar el electrodo en calibración o medición, asegúrese de: 1. Al desempacar, revise el electrodo para identificar cualquier daño mecánico en la membrana, diafragma o cuerpo. 2. Destornille primero el tapón, luego la tapa, y enjuague el electrodo con agua pura. Limpie el electrodo con un pañuelo. No frote la membrana de vidrio, ya que se puede aumentar el tiempo de respuesta. 3. Elimine cualquier burbuja dentro de la membrana. Equilibre la presión interna de la membrana con la de la atmósfera abriendo el orificio de relleno del electrodo al hacer una medición. Preguntas frecuentes: 1. ¿Qué significa el ícono de condición del electrodo? Después de la calibración, el medidor de pH determinará la pendiente y desfase del electrodo, y luego mostrará el resultado en la pantalla: Pendiente > 95% La condición del electrodo es muy buena Pendiente: 90%-95% La condición del electrodo es aceptable Pendiente: 85%-90% La condición del electrodo no es buena; necesita recalibrarse o reemplazarse. ST350 ES-3 2. ¿Cómo se almacena adecuadamente el electrodo de pH después de cada uso? R: Enjuague el electrodo con agua pura, coloque el electrodo en el tapón de humedecimiento. Si la solución en el tapón está contaminada o seca, límpielo y llénelo con solución 3M KCl. 3. ¿Cuál es la vida útil de los electrodos de pH? R: Las muestras medidas, las condiciones de almacenamiento y los métodos de mantenimiento influyen en la vida útil de los electrodos de pH; normalmente un electrodo de pH caduca después de 12 meses. Algunas muestras, como los ácidos o álcalis fuertes y los líquidos corrosivos acortan la vida útil del electrodo. El electrodo de pH es un consumible y generalmente se proporciona garantía. 4. ¿Por qué el valor de pH no es estable cuando se mide agua pura? R: Generalmente, el electrodo de pH común no puede medir agua destilada, agua de lluvia o agua pura. Esto se debe a que son líquidos y poseen baja conductividad. ST350 Modèle ST350 FR-1 Article n° 30129354 Description Électrode de pH 3 en 1 Spécifications : Connexion : BNC, Cinch Longueur du câble : 1 m Longueur de l'axe : 120 mm Diamètre de l'axe : 12 mm Gamme de pH : 0-14pH Potentiel zéro : 7,00 ±0,25 pH Pente : ≥ 97 % Impédance : ≤ 100 MΩ Composition de l'axe : Verre Électrolyte de référence : Solution 3M KCl Plage de température : 0-100℃ Sonde de température : NTC 30kΩ Diaphragme de référence : Manchon OHAUS offre différentes électrodes de pH pour différentes applications. S'assurer d'avoir choisi l'électrode de pH la mieux adaptée à vos échantillons ; procéder d'abord à un étalonnage de l'électrode. Les électrodes 3 en 1 contiennent une sonde de température. FR-2 ST350 Avant d'utiliser l'électrode pour un étalonnage ou une mesure, ce qui suit doit être noté : 1. Lors du déballage, vérifier l'électrode à la recherche d'endommagement mécanique de la membrane, du diaphragme ou du boîtier. 2. Dévisser d'abord le bouchon de mouillage, puis le couvercle et rincer l'électrode à l'eau pure. Essuyer l'électrode avec un papiermouchoir. Ne pas frotter la membrane de verre, sinon le temps de réponse peut augmenter. 3. Éliminer toutes les bulles d'air à l'intérieur de la membrane. Procéder à l'équilibre de la pression entre l'intérieur de l'électrode et l'atmosphère en ouvrant l'ouverture de remplissage de l'électrode lors de la mesure. Questions fréquentes : 1. Quelle est la signification de l'icône d'état de l'électrode ? Après l'étalonnage, le pH-mètre jugera la pente et le décalage de l'électrode avant d'afficher son jugement sur l'écran : pente > 95 % L'état de l'électrode est très bon pente : 90 %-95 % L'état de l'électrode est acceptable pente : 85 %-90 % L'état de l'électrode n'est pas bon ou nécessite un nettoyage ST350 FR-3 2. Comment pouvoir stocker correctement l'électrode de pH après chaque utilisation ? R : Rincer l'électrode à l'eau pure, ranger l'électrode dans le bouchon de mouillage. Si la solution dans le bouchon est polluée ou sèche, le nettoyer et le remplir avec une solution 3M KCl. 3. Quelle est la durée de vie d'une électrode de température ? R : Les échantillons mesurés, les conditions de stockage et les méthodes de maintenance peuvent tous influencer la durée de vie des électrodes de pH ; normalement, une électrode de pH devient périmée après 12 mois. Certains échantillons, comme des liquides corrosifs, fortement acides ou alcalins réduiront la durée de vie d'une électrode de pH. Une électrode de pH s'apparente à un consommable et est généralement fournie avec une garantie. 4. Pourquoi la valeur du pH est-elle instable lors de la mesure d'une eau pure ? R: Généralement, une électrode classique de pH ne peut pas mesurer l'eau distillée, l'eau de pluie ou l'eau pure. Parce que ce sont des liquides possédant une faible conductivité.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13

Ohaus ST350 Manual de usuario

Tipo
Manual de usuario

en otros idiomas