ResMed AirFit F30i Full Face Mask Manual de usuario

Tipo
Manual de usuario

Este manual también es adecuado para

English
Français
Español
Português
User guide
Device setting
Full face
USER GUIDE
English 1
ENGLISH
Thank you for choosing the AirFit
F30i full face mask. This mask uses a
cushion that seals under the nose and around the mouth, and a frame
with a tube connection above the head. To breathe fresh air, this mask
features vent holes in the cushion and elbow, and valves in the frame.
A
Elbow
4
Elbow ring
B
Frame
5
Anti-Asphyxia valve (inside the left
and right of frame connector)
C
Cushion
D
Headgear
6
Frame connector
1
Vent holes (cushion)
7
Magnetic clip
2
Vent holes (elbow)
8
Lower headgear strap
3
Side button
9
Upper headgear strap
Intended Use
The AirFit F30i mask is intended to be used by patients weighing more
than 66 lb (30 kg) who have been prescribed non-invasive positive airway
pressure (PAP) therapy such as CPAP or bi-level therapy. The mask is
intended for single patient re-use in the home and multi-patient re-use in
the hospital/institutional environment.
2
Contraindications
Use of masks with magnetic components is contraindicated in patients
with the following pre-existing conditions:
a metallic hemostatic clip implanted in your head to repair an
aneurysm
metallic splinters in one or both eyes.
Before using your mask
Remove all packaging and inspect each mask component for visible
deterioration.
Fitting your mask
English 3
1. Connect the air tubing from your device to the elbow.
2. Squeeze the side buttons on the elbow and pull from the frame. Put
the elbow and air tubing aside for now.
3. Twist and pull both magnetic clips away from the frame connectors.
4. Place the cushion under your nose and ensure it sits comfortably
against your face. With the ResMed logo on the headgear facing up,
pull the headgear and frame over your head.
5. Bring the lower headgear straps under your ears and attach the
magnetic clips to the frame connectors.
6. Undo the fastening tabs on the upper headgear straps and pull evenly.
Repeat with the lower headgear straps.
7. Attach the elbow to the top of the frame. Your mask should now be
positioned as shown.
Adjusting your mask
With the device turned on and blowing air, adjust the position of the
cushion for the most comfortable seal under your nose. Ensure that
the cushion is not creased and the headgear is not twisted.
To fix any mask leaks, adjust the upper or lower headgear straps.
Adjust only enough for a comfortable seal and do not overtighten.
Sizing the mask
Use the fitting template to assist in selecting the right size cushion.
If the mask frame falls back over your head or sits too close to your ears,
try a smaller frame size. If the mask falls forward on your head or sits too
close to your eyes, try a bigger frame size.
4
Disassembling your mask for cleaning
If your mask is connected to a device, disconnect the device air tubing
from the mask elbow.
1. Undo the fastening tabs on the upper headgear straps and pull from
the frame. Keep the magnetic clips attached to the lower headgear
straps.
2. Squeeze the side buttons on the elbow and detach from the frame.
3. Grip the frame connector and lift to unclip from the cushion. Repeat
on the other side.
Cleaning your mask
If there is any visible deterioration of a mask component (cracking,
crazing, tears etc.) the component should be discarded and replaced.
Daily/After each use: Cushion
Weekly: Headgear, frame and elbow
1. Soak the components in warm water with a mild liquid detergent.
2. Hand wash the components with a soft bristle brush. Pay particular
attention to the vent holes in both cushion and elbow.
3. Thoroughly rinse the components under running water.
4. Leave the components to air dry out of direct sunlight.
If the mask components are not visibly clean, repeat the cleaning steps.
Make sure that the vent holes and anti-asphyxia valves are clean and
clear.
English 5
Reassembling your mask
1. Align and insert the frame connector tab (a) into the cushion slot (b)
and press down until it clicks. Repeat on the other side.
2. Attach the elbow to the top of the frame until it clicks.
3. With the ResMed logo facing outwards and up, insert the upper
headgear straps into the frame from the inside and fold the fastening
tabs over to secure.
Note: If the elbow ring detaches, re-insert into the top of the frame.
WARNING
Magnets are used in the lower headgear straps and the frame of the mask. Ensure the
headgear and frame is kept at least 2" (50 mm) away from any active medical implant
(eg, pacemaker or defibrillator) to avoid possible effects from localized magnetic
fields. The magnetic field strength is less than 400 mT.
The mask contains safety features, the exhaust vent holes and anti-asphyxia valves, to
enable normal breathing and exhaust exhaled breath. Occlusion of the exhaust vent
holes or anti-asphyxia valves needs to be prevented to avoid having an adverse effect
on the safety and quality of the therapy. Regularly inspect the vent holes and anti-
asphyxia valves to ensure they are kept clean, clear of blockages and are not
damaged.
Only use compatible CPAP or bi-level therapy devices or accessories. The technical
specifications of the mask are provided for healthcare professionals to determine
compatible devices. Use in combination with incompatible medical devices can
decrease the safety or alter the performance of the mask.
The mask is not suitable for patients requiring life support ventilation.
Regularly clean your mask and its components to maintain the quality of your mask
and to prevent the growth of germs that can adversely affect your health.
6
WARNING
The mask must be used under qualified supervision for patients who are unable to
remove the mask by themselves. The mask may not be suitable for those prone to
aspiration.
The mask should not be used unless the device is turned on. Once the mask is fitted,
ensure the device is blowing air to reduce risk of rebreathing exhaled air.
Discontinue using or replace this mask if the patient has ANY adverse reaction to the
use of the mask. Consult your physician or sleep therapist.
The mask is not intended to be used simultaneously with nebulizer medications that
are in the air path of the mask/tube.
If there is any visible deterioration of a mask component (cracking, crazing, tears etc.)
the component should be discarded and replaced.
CAUTION
Always follow cleaning instructions and only use a mild liquid detergent. Some
cleaning products may damage the mask, its parts and their function, or leave harmful
residual vapours. Do not use a dish washer or washing machine to clean the mask.
When fitting the mask, do not overtighten the headgear as this may lead to skin
redness or sores around the mask cushion.
As with all masks, some rebreathing may occur at low pressures.
Using a mask may cause tooth, gum or jaw soreness or aggravate an existing dental
condition. If symptoms occur, consult your physician or dentist.
Do not iron the headgear as the material is heat sensitive and will be damaged.
English 7
Technical specifications
Mask setting options: For AirSense, AirCurve or S9: Select 'Full Face'.
Compatible devices: For a full list of compatible devices for this mask, see the
Mask/Device Compatibility List at ResMed.com/downloads/masks.
SmartStop may not operate effectively when using this mask with some CPAP or bi-level
devices.
Pressure-flow curve
Flow
(L/min)
22
38
48
59
67
Therapy pressure: 4 to 30 cm H
2
0
Resistance with Anti Asphyxia Valve closed to atmosphere
Drop in pressure measured (nominal) at 50 L/min: 0.2 cm H
2
O
Drop in pressure measured (nominal) at 100 L/min: 1.0 cm H
2
O
Resistance with Anti Asphyxia Valve open to atmosphere
Inspiration at 50 L/min: 0.2 cm H
2
O
Expiration at 50 L/min: 0.4 cm H
2
O
Anti-Asphyxia Valve open-to-atmosphere pressure: <4 cm H
2
O
Anti-Asphyxia Valve closed-to-atmosphere pressure: <4 cm H
2
O
Sound: Declared dual-number noise emission values in accordance with ISO4871:1996
and ISO3744:2010. A-weighted sound power level is 25 dBA, with uncertainty of 3 dBA. A-
weighted pressure level at a distance of 1 m is 18 dBA, with uncertainly of 3 dBA.
Environmental conditions
Operating temperature: 41°F to 104°F (5°C to 40°C)
Operating humidity: 15% to 95% RH non-condensing
Storage and transport temperature: -4°F to 140°F (-20°C to +60°C)
Storage and transport humidity: up to 95% RH non-condensing
International Commission on Non-Ionizing Radiation Protection (ICNIRP)
Magnets used in this mask are within ICNIRP guidelines for general public use.
Service life: The service life of the mask system is dependent on the intensity of usage,
maintenance, and environmental conditions to which the mask is used or stored. As this
mask system and its components are modular in nature, it is recommended that the user
maintain and inspect it on a regular basis, and replace the mask system or any
8
components if deemed necessary or according to the instructions in the ‘Cleaning your
mask’ section of this guide.
Accessories: Soft sleeves are available as an accessory,
Storage
Ensure that the mask is thoroughly clean and dry before storing it for any length of time.
Store the mask in a dry place out of direct sunlight.
Disposal
This mask does not contain any hazardous substances and may be disposed of with your
normal household refuse.
Reprocessing the mask between patients
Only Sleep Lab Mask (SLM) variants of the mask system are intended for multi-patient re-
use. When using between patients, these masks must be reprocessed according to
instructions available on ResMed.com/downloads/masks.
Symbols
The following symbols may appear on your product or packaging:
Full face mask
Device setting- Full face
Small frame
Standard frame
Large frame
Cushion size - small
Cushion size - medium
Cushion size - small wide
Cushion size - wide
Caution, consult accompanying documents.
Prescription only (In the US, Federal law restricts these devices to sale by or on
the order of a physician.)
See symbols glossary at ResMed.com/symbols.
English 9
Limited warranty
ResMed Ltd (hereafter ’ResMed’) warrants that your ResMed mask system (including mask
frame, cushion, headgear and tubing) shall be free from defects in material and
workmanship from the date of purchase for the period of 90 days or in the case of
disposable masks the maximum use period. This warranty is only available to the initial
consumer. It is not transferable. If the product fails under conditions of normal use,
ResMed will repair or replace, at its option, the defective product or any of its
components. This limited warranty does not cover: a) any damage caused as a result of
improper use, abuse, modification or alteration of the product; b) repairs carried out by any
service organization that has not been expressly authorized by ResMed to perform such
repairs; and c) any damage or contamination due to cigarette, pipe, cigar or other smoke.
Warranty is void on product sold, or resold, outside the region of original purchase.
Warranty claims on defective product must be made by the initial consumer at the point of
purchase.
This warranty replaces all other expressed or implied warranties, including any implied
warranty of merchantability or fitness for a particular purpose. Some regions or states do
not allow limitations on how long an implied warranty lasts, so the above limitation may
not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed to
have resulted from the sale, installation or use of any ResMed product. Some regions or
states do not allow the exclusion or limitation of incidental or consequential damages, so
the above limitation may not apply to you. This warranty gives you specific legal rights,
and you may also have other rights which vary from region to region. For further
information on your warranty rights, contact your local ResMed dealer or ResMed office.
Français 1
FRANÇAIS
Merci d’avoir choisi le masque facial d’AirFit
F30i. Ce masque dispose
d’une bulle qui adhère pour une parfaite étanchéité sous le nez et autour
de la bouche, et d’un entourage rigide muni d’un raccord de tuyau au-
dessus de la tête. Pour respirer de l’air frais, ce masque est doté
d’orifices de ventilation dans la bulle et le coude, ainsi que de valves dans
l’entourage rigide.
A
Coude
4
Bague du coude
B
Entourage rigide
5
Valve anti-asphyxie (dans le
raccord de l’entourage rigide à
gauche et à droite)
C
Bulle
D
Harnais
6
Raccord de l’entourage rigide
1
Orifices de ventilation (bulle)
7
Clip magnétique
2
Orifices de ventilation (coude)
8
Sangle inférieure du harnais
3
Bouton latéral
9
Sangle supérieure du harnais
Usage prévu
Le masque AirFit F30i est prévu pour un usage par des patients pesant
plus de 30 kg (66 lb) pour qui un traitement par pression positive non
invasif, comme un traitement par PPC ou à deux niveaux de pression, a
été prescrit. Le masque est prévu pour un usage multiple par un seul
2
patient à domicile ou pour un usage multiple par plusieurs patients à
l’hôpital ou en milieu médical.
Contre-indications
L’utilisation de masques avec des composants magnétiques est contre-
indiquée chez les patients présentant les pathologies ou états pré-
existants suivants :
dispositif hémostatique métallique implanté dans le crâne suite à un
anévrisme;
présence d’éclats métalliques dans un œil ou les deux yeux.
Avant d’utiliser votre masque
Retirez tout l’emballage et inspectez chaque composant du masque afin
de détecter toute détérioration visible.
Ajustement de votre masque
Français 3
1. Raccordez le circuit respiratoire de votre appareil au coude.
2. Appuyez sur les boutons latéraux du coude et retirez ce dernier de
l’entourage rigide. Mettez le coude et le circuit respiratoire de côté
pour le moment.
3. Tournez les clips magnétiques et détachez-les des raccords de
l’entourage rigide.
4. Placez la bulle sous votre nez et assurez-vous qu’elle repose
confortablement sur votre visage. Avec le logo ResMed du harnais
dirigé vers le haut, faites passer le harnais et l’entourage rigide par-
dessus votre tête.
5. Ramenez les sangles inférieures du harnais sous vos oreilles, puis
fixez les clips magnétiques aux raccords de l’entourage rigide.
6. Détachez les bandes de fixation des sangles supérieures du harnais et
tirez dessus uniformément. Répétez l’opération avec les sangles
inférieures du harnais.
7. Raccordez le coude sur le dessus de l’entourage rigide. Votre masque
doit être à présent positionné comme illustré.
Ajustement de votre masque
Avec l’appareil sous tension et émettant un débit d’air, ajustez la
position de la bulle pour qu’elle adhère de manière confortable sous
votre nez en assurant la meilleure étanchéité possible. Vérifiez que la
bulle ne présente aucun pli et que le harnais n’est pas entortillé.
En cas de fuite au niveau du masque, ajustez les sangles supérieures
ou inférieures du harnais. Ne les serrez pas trop : ajustez-les
uniquement pour obtenir une étanchéité confortable.
4
Détermination de la taille du masque
Servez-vous du gabarit d’ajustement pour vous aider à choisir la bonne
taille de bulle.
Si l’entourage rigide du masque tombe vers l’arrière ou se trouve trop
près de vos oreilles, essayez un entourage rigide de taille plus petite. Si
l’entourage rigide du masque tombe vers l’avant ou se trouve trop près
de vos yeux, essayez un entourage rigide de taille plus grande.
Démontage de votre masque pour le nettoyage
Si votre masque est connecté à un appareil, débranchez le circuit
respiratoire du coude du masque.
1. Détachez les bandes de fixation des sangles supérieures du harnais et
retirez-les de l’entourage rigide. Laissez les clips magnétiques sur les
sangles inférieures du harnais.
2. Appuyez sur les boutons latéraux du coude afin de le séparer de
l’entourage rigide.
3. Tenez fermement le raccord de l’entourage rigide et soulevez-le pour
le déclipser de la bulle. Répétez l’opération de l’autre côté.
Français 5
Nettoyage de votre masque
En cas de détérioration visible d’un composant du masque (fissures,
fendillements, déchirures, etc.), ce composant doit être jeté et remplacé.
Tous les jours/Après chaque utilisation : Bulle
Chaque semaine : Harnais, entourage rigide et coude
1. Faites tremper les composants dans une eau tiède avec un détergent
liquide doux.
2. Nettoyez les composants à la main à l’aide d’une brosse à soies
souples. Veillez en particulier à bien nettoyer les orifices de ventilation
situés à la fois dans la bulle et dans le coude.
3. Rincez soigneusement les composants sous l’eau courante.
4. Laissez les composants sécher à l’abri de la lumière directe du soleil.
Si les composants du masque semblent toujours sales, répétez les
différentes étapes de nettoyage. Vérifiez que les orifices de ventilation et
les valves anti-asphyxie sont propres et nets.
Remontage de votre masque
1. Alignez et insérez la languette du raccord de l’entourage rigide (a) dans
la fente de la bulle (b) et appuyez dessus jusqu’à ce qu’il s’enclenche.
Répétez l’opération de l’autre côté.
2. Fixez le coude sur le dessus de l’entourage rigide jusqu’à ce qu’il
s’enclenche.
3. Avec le logo ResMed dirigé vers l’extérieur et le haut, insérez les
sangles supérieures du harnais dans l’entourage rigide, de l’intérieur
vers l’extérieur, puis repliez les bandes de fixation pour les fermer.
Remarque : Si la bague du coude se détache, insérez-la de nouveau dans
le dessus de l’entourage rigide.
6
AVERTISSEMENT
Des aimants sont utilisés dans les sangles inférieures du harnais et dans l’entourage
rigide du masque. Vérifiez que le harnais et l’entourage rigide sont maintenus à une
distance minimale de 50 mm (2") de tout implant médical actif (par ex. un pacemaker
ou un défibrillateur) afin d’éviter les effets possibles des champs magnétiques
localisés. La force du champ magnétique est inférieure à 400 mT.
Le masque est doté de dispositifs de sécurité les orifices de ventilation et les valves
anti-asphyxie afin de permettre à la personne de respirer normalement et d’évacuer
l’air expiré. L’obstruction des orifices de ventilation ou des valves anti-asphyxie doit
être à tout prix évitée pour ne pas affecter la sécurité et la qualité du traitement.
Inspectez régulièrement les orifices de ventilation et les valves anti-asphyxie pour
veiller à leur propreté, leur bon état et leur dégagement.
N’utilisez que des accessoires ou des appareils de traitement par PPC ou à deux
niveaux de pression compatibles. Les caractéristiques techniques du masque
accompagnent ce dernier pour que les professionnels de santé puissent déterminer les
appareils compatibles. Utiliser le masque avec des appareils médicaux incompatibles
peut compromettre la sécurité ou altérer son efficacité.
Le masque ne convient pas aux patients nécessitant une ventilation support de vie.
Nettoyez régulièrement votre masque et ses composants pour maintenir leur qualité et
éviter la prolifération de germes potentiellement nuisibles à votre santé.
Le masque doit être utilisé sous le contrôle d’une personne qualifiée lorsque le patient
n’est pas en mesure de l’enlever de lui-même. Le masque peut ne pas convenir aux
patients sujets aux aspirations.
Le masque ne doit être porté que si l’appareil est sous tension. Une fois le masque en
place, assurez-vous que l’appareil émet un débit d’air afin de réduire le risque de
réinhalation d’air expiré.
En cas de réaction indésirable QUELCONQUE au masque de la part du patient, cessez
de l’utiliser ou remplacez-le. Consultez votre médecin ou un spécialiste du sommeil.
Ce masque n’est pas destiné à être utilisé conjointement à des médicaments
normalement administrés par nébuliseur qui se trouvent dans tout composant du
passage de l’air du masque ou du tuyau.
En cas de détérioration visible d’un composant du masque (fissures, fendillements,
déchirures, etc.), ce composant doit être jeté et remplacé.
Français 7
ATTENTION
Veillez à toujours respecter les instructions de nettoyage et à n’utiliser qu’un
détergent doux. Certains produits de nettoyage risquent d’endommager le masque,
ses pièces et leur fonctionnement ou de laisser des vapeurs résiduelles nocives.
Ne
nettoyez pas le masque à la machine à laver ou au lave-vaisselle.
Lors de l’ajustement du masque, veillez à ne pas trop serrer le harnais; cela peut
mener à l’apparition de rougeurs et d’irritations cutanées autour de la bulle du
masque.
Comme c’est le cas avec tous les masques, une réinhalation peut se produire à une
pression basse.
L’utilisation d’un masque peut provoquer des douleurs dans les dents, les gencives ou
la mâchoire ou aggraver un problème dentaire existant. Si ces symptômes
apparaissent, consultez un médecin ou un dentiste.
Ne repassez pas le harnais au risque d’endommager le matériau avec lequel il est
fabriqué, celui-ci étant sensible à la chaleur.
Caractéristiques techniques
Fonction de sélection du type de masque : Pour AirSense, AirCurve ou S9 :
Sélectionnez « Facial ».
Appareils compatibles : Pour une liste complète des appareils compatibles avec ce
masque, veuillez consulter la liste de compatibilité appareil/masque sur le site
www.ResMed.com/downloads/masks.
La fonction SmartStop peut ne pas fonctionner correctement lorsque ce masque est utili
avec certains appareils de PPC ou d'aide inspiratoire avec PEP.
Courbe pression/débit :
Débit
(l/min)
22
38
48
59
67
Pression de traitement : 4 à 30 cm H
2
O
Résistance avec la valve anti-asphyxie fermée
Chute de pression mesurée (nominale) à 50 l/min : 0,2 cm H
2
O
Chute de pression mesurée (nominale) à 100 l/min : 1,0 cm H
2
O
8
Résistance avec la valve anti-asphyxie ouverte
Inspiration à 50 l/min : 0,2 cm H
2
O
Expiration à 50 l/min : 0,4 cm H
2
O
Pression avec la valve anti-asphyxie ouverte :4 cm H
2
O
Pression avec la valve anti-asphyxie fermée :4 cm H
2
O
Niveau sonore : Valeurs d’émissions sonores à deux chiffres déclarées conformément
aux normes ISO4871:1996 et ISO3744:2010. Le niveau de puissance acoustique pondéré A
est de 25 dBA avec une incertitude de 3 dBA. Le niveau de pression acoustique pondéré A
mesuré à une distance d’1 m est de 18 dBA avec une incertitude de 3 dBA.
Conditions ambiantes
Température de fonctionnement : de 41°F à 104°F (de 5°C à 40°C)
Humidité de fonctionnement : de 15 % à 95 % HR sans condensation
Température de stockage et de transport : de -4°F à 140°F (de -20°C à +60°C)
Humidité de stockage et de transport : jusqu’à 95 % HR sans condensation
Commission internationale de protection contre les rayonnements non ionisants
(CIPRNI)
Les éléments magnétiques de ce masque respectent les directives de la CIPRNI pour une
utilisation grand public.
Durée de vie : La durée de vie du masque dépend de son utilisation, de l’entretien et des
conditions ambiantes dans lesquelles le masque est utilisé et rangé. Compte tenu de la
nature modulaire du masque et de ses composants, il est recommandé à l’utilisateur
d’entretenir et d’inspecter régulièrement le masque et de le remplacer ou de remplacer
tout composant le cas échéant ou conformément aux instructions indiquées dans la
section « Nettoyage de votre masque » de ce guide.
Accessoires : Des gaines souples sont proposées comme accessoires.
Stockage
Veillez à ce que le masque soit complètement propre et sec avant de le ranger. Rangez le
masque dans un endroit sec à l’abri de la lumière directe du soleil.
Élimination
Ce masque ne contient aucune substance dangereuse et peut être jeté avec les ordures
ménagères.
Retraitement du masque entre les patients
Seules les versions SLM (Sleep Lab Mask) du masque sont destinées à un usage multiple
par plusieurs patients. Avant d’être utilisés par un autre patient, ces masques doivent être
retraités conformément aux consignes disponibles sur ResMed.com/downloads/masks.
Français 9
Symboles
Les symboles suivants peuvent figurer sur votre produit ou sur son emballage :
Masque facial
Réglage de l’appareil - Facial
Entourage rigide de petite taille
Entourage rigide de taille standard
Entourage rigide de grande taille
Taille de la bulle - petite
Taille de la bulle - moyenne
Taille de la bulle - petite-large
Taille de la bulle - large
N'est pas fabriqué avec du latex
de caoutchouc naturel
Attention, veuillez consulter la documentation jointe.
Sur ordonnance uniquement (selon la loi fédérale américaine, cet appareil ne
peut être vendu aux États-Unis que par un médecin ou sur prescription médicale).
Reportez-vous au glossaire des symboles au ResMed.com/symbols.
Garantie limitée
ResMed Ltd (ci-après « ResMed ») garantit votre masque ResMed (y
compris l’entourage rigide, la bulle, le harnais et le circuit respiratoire)
contre tout défaut de matériaux et de main-d’œuvre pour une période de
90 jours à compter de la date d’achat ou, dans le cas d’un masque
jetable, pour la période d’utilisation maximale. Seul le client initial est
couvert par la présente garantie. Celle-ci n’est pas cessible. En cas de
défaillance du produit dans des conditions normales d’utilisation,
ResMed, à son entière discrétion, répare ou remplace le produit
défectueux ou toute pièce. La présente garantie limitée ne couvre pas : a)
tout dommage résultant d’une utilisation incorrecte, d’un usage abusif ou
d’une modification ou transformation opérée sur le produit; b) les
réparations effectuées par tout service de réparation sans l’autorisation
expresse de ResMed; et c) tout dommage ou contamination causé par de
la fumée de cigarette, de pipe, de cigare ou autre. La garantie est annulée
pour les produits vendus ou revendus dans un pays autre que celui où ils
ont été achetés à l’origine.
10
Les réclamations au titre de la garantie pour les produits défectueux
doivent être présentées au lieu d’achat par le client initial.
La présente garantie remplace toute autre garantie expresse ou implicite,
y compris toute garantie implicite de qualité marchande ou d’adéquation à
un usage particulier. Certains pays ou États n’autorisent pas les
limitations de durée pour les garanties implicites; il est donc possible que
la limitation susmentionnée ne s’applique pas à votre cas.
La société ResMed ne peut être tenue responsable de tout dommage
accessoire ou indirect résultant de la vente, de l’installation ou de
l’utilisation de tout produit ResMed. Certains pays ou États n’autorisent ni
l’exclusion ni la limitation des dommages accessoires ou indirects; il est
donc possible que la limitation susmentionnée ne s’applique pas à votre
cas. La présente garantie vous octroie des droits reconnus par la loi. Vous
pouvez également détenir d’autres droits qui varient en fonction du pays
où vous habitez. Pour de plus amples informations sur vos droits de
garantie, veuillez contacter ResMed ou votre revendeur local ResMed.
Español 1
ESPAÑOL
Gracias por elegir la mascarilla facial AirFit
F30i. En esta mascarilla se
utiliza una almohadilla que permite sellar debajo de la nariz y alrededor de
la boca, y un armazón que se conecta con el tubo por encima de la
cabeza. Para que se respire aire puro, esta mascarilla tiene orificios de
ventilación en la almohadilla y el codo, así como válvulas en el armazón.
A
Codo
4
Aro del codo
B
Armazón
5
Válvula antiasfixia (dentro del
lado izquierdo y derecho de la
conexión del armazón)
C
Almohadilla
D
Arnés
6
Conexión del armazón
1
Orificios de ventilación
(almohadilla)
7
Broche magnético
2
Orificios de ventilación (codo)
8
Correa inferior del arnés
3
Botón lateral
9
Correa superior del arnés
Uso indicado
La mascarilla AirFit F30i está diseñada para ser utilizada en pacientes que
pesen más de 30 kg (66 lb) a los que se les haya indicado tratamiento no
invasivo con presión positiva en las vías respiratorias (PAP), como un
tratamiento CPAP o binivel. La mascarilla está diseñada para ser utilizada
en repetidas ocasiones por un único paciente en el domicilio, y por varios
pacientes en hospitales o instituciones.
2
Contraindicaciones
El uso de las mascarillas con componentes magnéticos está
contraindicado en pacientes que tengan las siguientes afecciones
preexistentes:
una pinza hemostática de metal implantada en la cabeza para reparar
un aneurisma;
esquirlas metálicas en uno o ambos ojos.
Antes de utilizar la mascarilla
Quite todo el embalaje e inspeccione cada una de las piezas para
asegurarse de que no haya signos visibles de deterioro.
Colocación de la mascarilla
Español 3
1. Conecte el tubo de aire del equipo al codo.
2. Apriete los botones laterales del codo y tire para separarlo del
armazón. Deje el codo y el tubo de aire a un lado por ahora.
3. Gire ambos broches magnéticos y tire de ellos para separarlos de las
conexiones del armazón.
4. Colóquese la almohadilla debajo de la nariz y asegúrese de que quede
cómodamente apoyada sobre el rostro. Coloque el arnés con el
logotipo de ResMed mirando hacia arriba y pase el arnés y el armazón
por encima de la cabeza.
5. Pase las correas inferiores del arnés por debajo de las orejas y
conecte los broches magnéticos a las conexiones del armazón.
6. Despegue las lengüetas de sujeción de las correas superiores del
arnés y tire de forma pareja. Repita ese paso con las correas inferiores
del arnés.
7. Conecte el codo a la parte superior del armazón. La mascarilla debe
quedar colocada como se muestra en la figura.
Ajuste de la mascarilla
Con el equipo encendido y suministrando aire, corrija la posición de la
almohadilla para lograr el sellado que le resulte más cómodo debajo de
la nariz. Asegúrese de que la almohadilla no esté plegada y el arnés no
esté torcido.
Para resolver las fugas que pueda haber por la mascarilla, ajuste las
correas superiores o inferiores del arnés. Ajústelas lo que sea
necesario para obtener un sellado cómodo: no las apriete demasiado.
4
Selección del tamaño de la mascarilla
Use el modelo de ajuste para seleccionar una almohadilla que tenga el
tamaño correcto.
Si el armazón de la mascarilla se le desliza hacia atrás por la cabeza o le
queda apoyado muy cerca de las orejas, pruebe con uno de tamaño más
pequeño. Si la mascarilla se le desliza hacia adelante por la cabeza o le
queda apoyada muy cerca de los ojos, pruebe con un armazón más
grande.
Desmontaje de la mascarilla para limpiarla
Si la mascarilla está conectada a un equipo, tome el tubo de aire de este
último y desconéctelo del codo de la mascarilla.
1. Despegue las lengüetas de sujeción de las correas superiores del
arnés y tire de ellas para separarlas del armazón. Mantenga los
broches magnéticos unidos a las correas inferiores del arnés.
2. Apriete los botones laterales ubicados en el codo y desconéctelo del
armazón.
3. Tome la conexión del armazón y levántela para desabrocharla de la
almohadilla. Haga lo mismo en el otro lado.
Español 5
Limpieza de la mascarilla
Si advierte algún signo de deterioro en alguna de las piezas de la
mascarilla (como grietas, rajaduras, roturas, etc.), la pieza debe
desecharse y sustituirse por otra.
Diariamente/después de cada uso: almohadilla
Semanal: arnés, armazón y codo.
1. Sumerja las piezas en agua tibia con detergente líquido suave.
2. Lave las piezas a mano con un cepillo de cerdas suaves. Preste
especial atención a los orificios de ventilación tanto de la almohadilla
como del codo.
3. Enjuague las piezas por completo bajo agua de grifo.
4. Deje secar las piezas al aire, sin exponerlas a la luz solar directa.
Si las piezas de la mascarilla no se ven limpias, repita los pasos de
limpieza. Compruebe que los orificios de ventilación y las válvulas
antiasfixia estén limpios y despejados.
Montaje de la mascarilla
1. Coloque la lengüeta de la conexión del armazón (a) a la misma altura
que la ranura de la almohadilla (b), insértela en ella y presione hacia
abajo hasta que haga clic. Haga lo mismo en el otro lado.
2. Conecte el codo en la parte superior del armazón hasta que haga clic.
3. Con el logotipo de ResMed mirando hacia afuera y hacia arriba, inserte
las correas superiores del arnés en el armazón desde el lado interior y
doble las lengüetas de sujeción para sujetar.
Nota: Si el aro del codo se desconecta, vuelva a insertarlo en la parte
superior del armazón.
6
ADVERTENCIA
Se utilizan imanes en las correas inferiores del arnés y en el armazón de la mascarilla.
Asegúrese de que el arnés y el armazón se mantengan por lo menos a 50 mm (2 pulg.)
de distancia de cualquier implante médico activo (p. ej. un marcapaso o desfibrilador),
para evitar posibles efectos provocados por los campos magnéticos localizados. La
intensidad de los campos magnéticos es inferior a 400 mT.
La mascarilla tiene características de seguridad los orificios de ventilación y las
válvulas antiasfixia, para que se pueda respirar normalmente y expulsar el aire
exhalado. Es necesario evitar que los orificios de ventilación o las válvulas antiasfixia
se obstruyan, para no perjudicar la seguridad y la calidad del tratamiento. Inspeccione
con regularidad los orificios de ventilación y las válvulas antiasfixia para cerciorarse
de que estén limpios, no tengan obstrucciones y no estén dañados.
Utilice solo accesorios o equipos de tratamiento binivel o CPAP que sean compatibles.
Las especificaciones técnicas de la mascarilla se proporcionan para que los
profesionales de salud puedan determinar qué equipos son compatibles. Usarla con
equipos médicos incompatibles puede disminuir la seguridad o alterar el rendimiento
de la mascarilla.
La mascarilla no es adecuada para utilizarla en pacientes que necesitan ventilación de
apoyo vital.
Limpie la mascarilla y sus piezas con regularidad para preservar la calidad de la
mascarilla y evitar la presencia de microbios que pueden ser perjudiciales para la
salud.
La mascarilla debe ser utilizada bajo supervisión calificada cuando los pacientes no
sean capaces de retirarla por sí mismos. Es posible que la mascarilla no sea apropiada
para quienes sean propensos a la aspiración.
La mascarilla no debe utilizarse a menos que el equipo se encuentre encendido. Una
vez que la mascarilla esté colocada, compruebe que el equipo esté suministrando aire
para reducir el riesgo de volver a inspirar el aire exhalado.
Deje de usar esta mascarilla o reemplácela si el paciente presenta CUALQUIER
reacción adversa a su utilización. Consulte al médico o al especialista en medicina del
sueño.
No se debe utilizar la mascarilla junto con medicamentos para nebulizadores ubicados
en el circuito de aire de la mascarilla o el tubo.
Si advierte algún signo de deterioro en alguna de las piezas de la mascarilla (como
grietas, rajaduras, roturas, etc.), la pieza debe desecharse y sustituirse por otra.
Español 7
PRECAUCIÓN
Siga siempre las instrucciones de limpieza y utilice únicamente un detergente líquido
suave. Algunos productos de limpieza pueden dañar la mascarilla, sus piezas y su
funcionamiento, o dejar vapores residuales dañinos.
No lave la mascarilla en un
lavavajillas o lavarropas.
Cuando coloque la mascarilla, no apriete demasiado el arnés, ya que esto puede
provocar enrojecimiento de la piel o llagas alrededor de la almohadilla.
Como ocurre con todas las mascarillas, si la presión es baja se puede volver a inhalar
parte del aire exhalado.
Utilizar una mascarilla puede provocar dolor en los dientes o muelas, en las encías o
en la mandíbula, o agravar una afección dental ya existente. Si se presentan síntomas,
consulte al médico o al odontólogo.
No planche el arnés, ya que está hecho de un material sensible al calor y se dañará.
Especificaciones técnicas
Opciones de configuración de la mascarilla: en el AirSense, AirCurve o S9, seleccione
Facial.
Equipos compatibles: para obtener una lista completa de los equipos compatibles con
esta mascarilla, consulte la lista de compatibilidad entre mascarillas y equipos en
ResMed.com/downloads/masks.
Es posible que SmartStop no funcione eficazmente al usar esta mascarilla con algunos
equipos CPAP o binivel.
Curva de flujo en función de la presión
Presión
(cm H
2
O)
Flujo
(l/min)
4
22
11
38
17
48
24
59
30
67
Presión de tratamiento: de 4 a 30 cm H
2
O
Resistencia con válvula antiasfixia cerrada a la atmósfera
Caída de presión medida (nominal) a 50 l/min: 0,2 cm H
2
O
Caída de presión medida (nominal) a 100 l/min: 1,0 cm H
2
O
Resistencia con válvula antiasfixia abierta a la atmósfera
Inspiración a 50 l/min: 0,2 cm H
2
O
Espiración a 50 l/min: 0,4 cm H
2
O
8
Presión con válvula antiasfixia abierta a la atmósfera: <4 cm H
2
O
Presión con válvula antiasfixia cerrada a la atmósfera: <4 cm H
2
O
Sonido: valores de emisión sonora declarados mediante dos números de conformidad con
la ISO4871:1996 y la ISO3744:2010. El nivel de potencia acústica ponderado según la
escala A es de 25 dBA, con una incertidumbre de 3 dBA. El nivel de presión acústica
ponderado según la escala A a una distancia de 1 m es de 18 dBA, con una incertidumbre
de 3 dBA.
Condiciones ambientales
Temperatura de funcionamiento: de 5 °C a 40 °C (de 41 °F a 104 °F)
Humedad de funcionamiento: de 15 % a 95 % de HR sin condensación
Temperatura de almacenamiento y transporte: de -20 °C a +60 °C (de -4 °F a 140 °F)
Humedad de almacenamiento y transporte: hasta 95 % de HR sin condensación
Comisión Internacional de Protección contra la Radiación No Ionizante (ICNIRP)
Los imanes que se usan en esta mascarilla cumplen con las directrices de la ICNIRP para
uso por parte del público en general.
Vida útil: la vida útil del sistema de la mascarilla depende de la intensidad con que se lo
use, el mantenimiento que se le dé y las condiciones ambientales en que la mascarilla se
use o almacene. Dado que este sistema de mascarilla y sus piezas son de naturaleza
modular, se recomienda al usuario que le dé mantenimiento y lo inspeccione con
regularidad, y que cambie el sistema o cualquiera de sus piezas si lo considera necesario o
si así se indica en las instrucciones que se brindan en la sección “Limpieza de la
mascarilla” del presente manual.
Accesorios: hay fundas blandas disponibles como accesorio.
Almacenamiento
Asegúrese de que la mascarilla esté completamente limpia y seca antes de guardarla
durante cualquier período considerable de tiempo. Guarde la mascarilla en un lugar seco,
fuera de la luz solar directa.
Eliminación
La mascarilla no contiene ninguna sustancia peligrosa y puede eliminarse con los residuos
domésticos habituales.
Reprocesamiento de la mascarilla de un paciente a
otro
Solo las variantes SLM (para usar en laboratorios del sueño) del sistema de la mascarilla
están diseñadas para utilizarse en más de un paciente. Cuando se vaya a cambiar de
paciente, estas mascarillas se deben reprocesar de conformidad con las instrucciones que
se brindan en ResMed.com/downloads/masks.
Español 9
Símbolos
Los símbolos que figuran a continuación pueden aparecer en el producto o en el envoltorio:
Mascarilla facial
Configuración del equipo: facial
Armazón pequeño
Armazón estándar
Armazón grande
Tamaño de la almohadilla:
pequeño
Tamaño de la almohadilla:
mediano
Tamaño de la almohadilla: ancho y
pequeño
Tamaño de la almohadilla: ancho
No hecho con látex de caucho
natural
Precaución: consulte los
documentos adjuntos.
Solo con receta (en EE. UU., la ley federal exige que estos equipos sean vendidos
únicamente por un médico o por su orden)
Véase el glosario de símbolos en ResMed.com/symbols.
Garantía limitada
ResMed Ltd (en adelante “ResMed”) garantiza que su sistema de
mascarilla ResMed (incluidos el armazón, la almohadilla, el arnés y el tubo
de la mascarilla) estará libre de todo defecto de material y mano de obra a
partir de la fecha de adquisición y durante un período de 90 días o, en el
caso de las mascarillas desechables, durante el período máximo de uso.
Esta garantía solo tiene validez para el consumidor inicial y no es
transferible. Si el producto falla en condiciones de utilización normales,
ResMed reparará o reemplazará, a opción de ResMed, el producto
defectuoso o cualquiera de sus piezas. Esta garantía limitada no cubre: a)
ningún daño provocado por la utilización indebida, el abuso, la
modificación o la alteración del producto; b) reparaciones llevadas a cabo
por cualquier organización dedicada a la reparación que no haya sido
expresamente autorizada por ResMed para hacer dichas reparaciones, y
c) ningún daño o contaminación provocados por humo de cigarrillo, pipa,
habano u otros tipos de humo. La garantía queda anulada si el producto
se vende o revende fuera de la región de compra original.
10
Las reclamaciones de garantía con respecto a productos defectuosos
deben ser realizadas por el consumidor original en el punto de compra.
Esta garantía reemplaza cualquier otra garantía expresa o implícita,
incluida cualquier garantía implícita de comerciabilidad o idoneidad para un
propósito en particular. Algunas regiones o estados no permiten
limitaciones respecto de la duración de una garantía implícita, por lo que
es posible que la limitación estipulada anteriormente no sea aplicable en
su caso particular.
ResMed no se responsabilizará de ningún daño incidental ni emergente
que se diga que ocurrió a consecuencia de la venta, la instalación o el uso
de alguno de los productos de ResMed. Algunas regiones o estados no
permiten la exclusión o la limitación de los daños incidentales o
emergentes, por lo que es posible que la limitación estipulada
anteriormente no sea aplicable en su caso particular. La presente garantía
le otorga derechos legales específicos, y es posible que usted también
tenga otros derechos que pueden variar de una región a otra. Para
obtener s información acerca de los derechos de garantía, póngase en
contacto con el distribuidor de ResMed o la sucursal de ResMed de su
zona.
Português 1
PORTUGUÊS
Obrigado por escolher a máscara oronasal AirFit
F30i. Essa máscara
possui uma almofada que veda a região abaixo do nariz e ao redor da
boca, além de uma armação com uma conexão de tubo acima da cabeça.
Para respirar ar fresco, a máscara conta com orifícios do respiradouro na
almofada e no cotovelo, e válvulas na armação.
A
Cotovelo
4
Anel do cotovelo
B
Armação
5
Válvula antiasfixia (dentro da
parte esquerda e direita do
conector da armação)
C
Almofada
D
Arnês
6
Conector da armação
1
Orifícios do respiradouro
(almofada)
7
Presilha magnética
2
Orifícios do respiradouro
(cotovelo)
8
Correia inferior do arnês
3
Botão lateral
9
Correia superior do arnês
Uso previsto
A máscara AirFit F30i foi projetada para pacientes com peso superior a
30 kg (66 lb) que receberam prescrição de terapia de pressão positiva não
invasiva nas vias respiratórias (PAP), como CPAP ou terapia de dois
níveis. A máscara foi projetada para reutilização por um único paciente em
2
ambiente doméstico e reutilização por vários pacientes em ambiente
hospitalar/institucional.
Contraindicações
O uso de máscaras com componentes magnéticos é contraindicado em
pacientes que já apresentem as seguintes condições:
clipes hemostáticos metálicos implantados na cabeça para reparação
de um aneurisma
fragmentos metálicos em um ou em ambos os olhos.
Antes de usar a máscara
Remova todas as embalagens e verifique se algum componente da
máscara apresenta deterioração visível.
Colocação da máscara
Português 3
1. Conecte o tubo de ar do dispositivo ao cotovelo.
2. Pressione os botões laterais no cotovelo e puxe a peça da armação.
Deixe o cotovelo e o tubo de ar de lado agora.
3. Gire e puxe as duas presilhas magnéticas para fora dos conectores da
armação.
4. Coloque a almofada sob o nariz e veja se está confortável em seu
rosto. Com o logotipo da ResMed no arnês voltado para cima, passe o
arnês e a armação sobre a cabeça.
5. Passe as correias inferiores do arnês por baixo das orelhas e prenda
as presilhas magnéticas aos conectores da armação.
6. Solte as pontas de fixação das correias superiores do arnês e puxe
uniformemente. Repita com as correias inferiores do arnês.
7. Prenda o cotovelo à parte de cima da armação. A máscara deve ser
posicionada conforme mostrado.
Ajuste da máscara
Com o dispositivo ligado e soprando ar, ajuste a posição da almofada
para que vede do modo mais confortável possível sob o nariz.
Certifique-se de que a almofada não esteja dobrada e o arnês não
esteja torcido.
Para consertar fugas da máscara, ajuste as correias inferiores e
superiores do arnês. Ajuste apenas o suficiente para uma vedação
confortável e não aperte demais.
Tamanho da máscara
Use o modelo de ajuste para auxiliar na seleção da almofada no tamanho
correto.
Se a armação da máscara ficar muito para trás da cabeça ou muito
próxima dos ouvidos, experimente um tamanho de armação menor. Se a
4
máscara ficar muito para a frente da cabeça ou muito próxima dos olhos,
experimente um tamanho de armação maior.
Desmontagem da máscara para limpeza
Se a máscara estiver conectada a um dispositivo, desconecte o tubo de
ar do dispositivo pelo cotovelo.
1. Solte as pontas de fixação nas correias superiores do arnês e puxe
para liberá-las da armação. Mantenha as presilhas magnéticas fixadas
nas correias inferiores do arnês.
2. Pressione os botões laterais no cotovelo e solte a peça da armação.
3. Aperte o conector da armação e levante para soltá-lo da almofada.
Repita do outro lado.
Português 5
Limpeza da máscara
Se for observado que algum componente da máscara se encontra
deteriorado (quebrado, rachado, roto, etc.), tal componente deve ser
descartado e substituído.
Diariamente/Após cada utilização: Almofada
Semanalmente: Arnês, armação e cotovelo
1. Mergulhe os componentes em água morna com um detergente
líquido suave.
2. Lave manualmente os componentes com uma escova de cerdas
macias. Preste muita atenção aos orifícios do respiradouro na
almofada e no cotovelo.
3. Enxágue completamente os componentes com água corrente.
4. Deixe os componentes secar sem contato com luz solar direta.
Se os componentes da máscara não estiverem visivelmente limpos,
repita as etapas de limpeza. Certifique-se de que os orifícios do
respiradouro e as válvulas antiasfixia estejam limpos e desobstruídos.
Remontagem da máscara
1. Alinhe e insira a aba do conector da armação (a) à fenda da almofada
(b) e pressione até ouvir um clique. Repita do outro lado.
2. Prenda o cotovelo à parte de cima da armação até ouvir um clique.
3. Com o logotipo da ResMed voltado para cima e para fora, insira as
correias superiores do arnês na armação de dentro para fora e dobre
as pontas de fixação para prender.
Nota: Se o anel do cotovelo se soltar, insira novamente na parte de cima
da armação.
6
AVISO
São utilizados ímãs nas correias inferiores do arnês e na armação da máscara.
Mantenha o arnês e a armação a no mínimo 50 mm (2 pol.) de distância de qualquer
implante médico ativo (por exemplo, marca-passo ou desfibrilador) para evitar os
possíveis efeitos de campos magnéticos localizados.
A intensidade do campo
magnético é inferior a 400 mT.
A máscara contém recursos de segurança, os orifícios do respiradouro e as válvulas
antiasfixia, para permitir que o paciente respire e exale o ar normalmente. Evite a
oclusão dos orifícios do respiradouro ou das válvulas antiasfixia para impedir a
ocorrência de efeitos adversos na segurança e qualidade do tratamento. Inspecione
frequentemente os orifícios do respiradouro e as válvulas antiasfixia para garantir que
eles estejam limpos, sem qualquer bloqueio e não estejam danificados.
Utilize apenas dispositivos e acessórios terapêuticos CPAP ou de dois níveis
compatíveis. As especificações técnicas da máscara são fornecidas para profissionais
de saúde para determinar dispositivos compatíveis. O uso em combinação com
dispositivos médicos incompatíveis pode diminuir a segurança ou alterar o
desempenho da máscara.
A máscara não é adequada para pacientes que necessitam de ventilação de suporte à
vida.
Limpe regularmente sua máscara e seus componentes para manter a qualidade de sua
máscara e para impedir a proliferação de germes que podem afetar negativamente
sua saúde.
Pacientes que não consigam retirar a máscara sozinhos devem usar a máscara sob a
supervisão de pessoas qualificadas. A máscara pode não ser adequada para pessoas
predispostas a aspiração.
A máscara só deve ser usada se o dispositivo estiver ligado. Depois de colocada a
máscara, certifique-se de que o dispositivo esteja soprando ar, para reduzir o risco de
reinalação do ar exalado.
Interrompa o uso ou substitua essa máscara, se o paciente tiver QUALQUER reação
adversa ao uso da máscara. Consulte seu médico ou terapeuta de sono.
A máscara não deve ser usada concomitantemente com medicamentos de nebulizador
situados na passagem de ar da máscara/do tubo.
Se for observado que algum componente da máscara se encontra deteriorado
(quebrado, rachado, roto, etc.), tal componente deve ser descartado e substituído.
Português 7
PRECAUÇÃO
Siga sempre as instruções de limpeza e use apenas detergente líquido suave. Alguns
produtos de limpeza podem danificar a máscara, suas peças e funções, ou deixar
vapores residuais nocivos. Não use máquinas de lavar louça ou roupa para limpar a
máscara.
Ao colocar a máscara, não aperte demais o arnês, pois isso pode causar vermelhidão
ou feridas na pele ao redor da almofada da máscara.
Como em todas as máscaras, poderá ocorrer uma certa reinalação em pressões
baixas.
O uso da máscara pode deixar os dentes, a gengiva ou a mandíbula doloridos, ou
agravar algum problema odontológico existente. Se aparecerem sintomas, consulte
seu médico ou dentista.
Não passe o arnês a ferro, pois o material é sensível ao calor e será danificado.
Especificações técnicas
Opções de configuração da máscara: Para AirSense, AirCurve ou S9: Selecione
"Facial".
Dispositivos compatíveis: Para obter uma lista completa de dispositivos compatíveis
com esta máscara, consulte a Lista de Compatibilidade entre Máscara/Dispositivo em
ResMed.com/downloads/masks.
O SmartStop poderá não funcionar efetivamente quando esta máscara for usada com
certos dispositivos CPAP ou de dois níveis.
Curva de fluxo/pressão
Pressão
(cm H
2
O)
Fluxo
(l/min)
4
22
11
38
17
48
24
59
30
67
Pressão terapêutica: 4 a 30 cm H
2
0
Resistência com válvula antiasfixia fechada para o ambiente
Medição da queda de pressão (nominal) a 50 l/min: 0,2 cm H
2
O
Medição da queda de pressão (nominal) a 100 l/min: 1,0 cm H
2
O
Resistência com válvula antiasfixia aberta para o ambiente
Inspiração a 50 l/min: 0,2 cm H
2
O
8
Expiração a 50 l/min: 0,4 cm H
2
O
Pressão com válvula antiasfixia aberta para o ambiente: <4 cm H
2
O
Pressão com válvula antiasfixia fechada para o ambiente: <4 cm H
2
O
Som: Valores declarados de emissão de ruído expressos por um número duplo em
conformidade com a ISO4871:1996 e ISO3744:2010. O nível de potência sonora ponderado
em A é de 25 dBA, com imprecisão de 3 dBA. O nível de pressão sonora ponderado em A a
uma distância de 1 m é de 18 dBA, com imprecisão de 3 dBA.
Condições ambientais
Temperatura operacional: 5°C a 40°C (41°F a 104°F)
Umidade operacional: 15% a 95% UR sem condensação
Temperatura de armazenamento e transporte: -20°C a +60°C (-4°F a 140°F)
Umidade de armazenamento e transporte: até 95% UR sem condensação
Comissão Internacional de Proteção Contra Radiação Não Ionizante (ICNIRP)
Os ímãs usados nesta máscara obedecem às diretrizes da ICNIRP para uso pelo público em
geral.
Vida útil: A vida útil do sistema de máscara depende da intensidade de uso, da
manutenção e das condições ambientais nas quais a máscara é utilizada ou armazenada.
Como o sistema de máscara e seus componentes têm natureza modular, é recomendável
que o usuário os conserve e examine regularmente e substitua o sistema ou qualquer
componente em caso de necessidade ou de acordo as instruções na seção “Limpeza da
máscara” deste guia.
Acessórios: Luvas maleáveis estão disponíveis como acessórios.
Armazenamento
Certifique-se de que a máscara esteja totalmente limpa e seca antes de guardá-la por
qualquer período de tempo. Guarde a máscara em local seco, ao abrigo da luz solar direta.
Eliminação
Esta máscara não contém qualquer substância perigosa e pode ser eliminada com o lixo
doméstico.
Reprocessamento da máscara entre pacientes
Somente as variantes SLM (Sleep Lab Mask) do sistema de máscara foram projetadas
para reutilização por vários pacientes. Em caso de uso em outros pacientes, estas
máscaras têm de ser reprocessadas de acordo com as instruções disponíveis em
ResMed.com/downloads/masks.
Português 9
Símbolos
Os símbolos a seguir podem aparecer no seu produto ou na embalagem:
Máscara oronasal
Configuração do dispositivo -
oronasal
Armação pequena
Armação padrão
Armação grande
Tamanho da almofada - pequena
Tamanho da almofada - média
Tamanho da almofada - pequena
larga
Tamanho da almofada - larga
Não é feito com látex de borracha
natural
Precaução, consulte os
documentos anexos.
Somente com prescrição (nos EUA, a lei federal limita a venda destes
dispositivos a médicos ou por prescrição médica).
Consulte o glossário de símbolos em ResMed.com/symbols.
Garantia limitada
A ResMed Ltd (daqui por diante, "ResMed") garante que seu sistema de
máscara ResMed (incluindo a armação da máscara, a almofada, o arnês e
os tubos) está livre de defeitos de material e mão de obra por um período
de 90 dias a partir da data de compra ou, no caso de máscaras
descartáveis, pelo período máximo de uso. Esta garantia só está
disponível para o consumidor inicial. Não é transferível. Se o produto
apresentar defeito em condições normais de utilização, a ResMed
efetuará, por opção própria, o reparo ou a substituição do produto
defeituoso ou de qualquer um de seus componentes. Esta garantia
limitada não cobre: a) danos provocados por uso incorreto, uso abusivo,
modificação ou alteração do produto; b) consertos efetuados por
empresas de assistência técnica que não tenham sido expressamente
autorizadas pela ResMed para efetuar tais serviços; e c) danos ou
contaminações provocadas por fumaça de cigarro, cachimbo, charuto ou
outras fontes. A garantia deixa de ser válida se o produto for vendido, ou
revendido, fora da região de compra original.
10
Os pedidos de reparo ou substituição de um produto defeituoso no
âmbito da garantia devem ser feitos pelo consumidor original no local de
compra.
Esta garantia substitui todas as outras, explícitas ou implícitas, incluindo
qualquer garantia implícita de comerciabilidade ou de adequabilidade para
determinado fim. Algumas regiões ou estados não permitem limitações
de tempo sobre a duração de uma garantia implícita, pelo que a limitação
acima pode não se aplicar ao seu caso.
A ResMed não será responsabilizada por quaisquer danos incidentais ou
consequentes reivindicados como decorrentes da venda, instalação ou
uso de qualquer produto ResMed. Algumas regiões ou estados não
permitem a exclusão ou limitação de danos incidentais ou consequentes;
assim, a limitação acima pode não se aplicar ao seu caso. Esta garantia
confere-lhe direitos legais específicos, e você pode ter outros direitos que
variam de região para região. Para obter mais informações sobre seus
direitos de garantia, entre em contato com o revendedor ou escritório
local da ResMed.
ResMed.com
ResMed Pty Ltd
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153 Australia
See ResMed.com for other ResMed locations worldwide. AirFit, AirSense, AirCurve and S9 are trademarks
and/or registered trademarks of the ResMed family of companies. For patent and other intellectual property
information, see ResMed.com/ip. © 2020 ResMed. 638225/1 2020-02
Device setting
Full face
en Full face mask
fr
Masque facial
es Mascarilla facial
pt Máscara oronasal
638225

Transcripción de documentos

English Français Español Português USER GUIDE Device setting Full face ENGLISH Thank you for choosing the AirFit™ F30i full face mask. This mask uses a cushion that seals under the nose and around the mouth, and a frame with a tube connection above the head. To breathe fresh air, this mask features vent holes in the cushion and elbow, and valves in the frame. A B Elbow Frame 4 5 C D 1 2 3 Cushion Headgear Vent holes (cushion) Vent holes (elbow) Side button Elbow ring Anti-Asphyxia valve (inside the left and right of frame connector) 6 7 8 9 Frame connector Magnetic clip Lower headgear strap Upper headgear strap Intended Use The AirFit F30i mask is intended to be used by patients weighing more than 66 lb (30 kg) who have been prescribed non-invasive positive airway pressure (PAP) therapy such as CPAP or bi-level therapy. The mask is intended for single patient re-use in the home and multi-patient re-use in the hospital/institutional environment. English 1 Contraindications Use of masks with magnetic components is contraindicated in patients with the following pre-existing conditions: • a metallic hemostatic clip implanted in your head to repair an aneurysm • metallic splinters in one or both eyes. Before using your mask Remove all packaging and inspect each mask component for visible deterioration. Fitting your mask 2 1. Connect the air tubing from your device to the elbow. 2. Squeeze the side buttons on the elbow and pull from the frame. Put the elbow and air tubing aside for now. 3. Twist and pull both magnetic clips away from the frame connectors. 4. Place the cushion under your nose and ensure it sits comfortably against your face. With the ResMed logo on the headgear facing up, pull the headgear and frame over your head. 5. Bring the lower headgear straps under your ears and attach the magnetic clips to the frame connectors. 6. Undo the fastening tabs on the upper headgear straps and pull evenly. Repeat with the lower headgear straps. 7. Attach the elbow to the top of the frame. Your mask should now be positioned as shown. Adjusting your mask • With the device turned on and blowing air, adjust the position of the cushion for the most comfortable seal under your nose. Ensure that the cushion is not creased and the headgear is not twisted. • To fix any mask leaks, adjust the upper or lower headgear straps. Adjust only enough for a comfortable seal and do not overtighten. Sizing the mask Use the fitting template to assist in selecting the right size cushion. If the mask frame falls back over your head or sits too close to your ears, try a smaller frame size. If the mask falls forward on your head or sits too close to your eyes, try a bigger frame size. English 3 Disassembling your mask for cleaning If your mask is connected to a device, disconnect the device air tubing from the mask elbow. 1. Undo the fastening tabs on the upper headgear straps and pull from the frame. Keep the magnetic clips attached to the lower headgear straps. 2. Squeeze the side buttons on the elbow and detach from the frame. 3. Grip the frame connector and lift to unclip from the cushion. Repeat on the other side. Cleaning your mask If there is any visible deterioration of a mask component (cracking, crazing, tears etc.) the component should be discarded and replaced. Daily/After each use: Cushion Weekly: Headgear, frame and elbow 1. Soak the components in warm water with a mild liquid detergent. 2. Hand wash the components with a soft bristle brush. Pay particular attention to the vent holes in both cushion and elbow. 3. Thoroughly rinse the components under running water. 4. Leave the components to air dry out of direct sunlight. If the mask components are not visibly clean, repeat the cleaning steps. Make sure that the vent holes and anti-asphyxia valves are clean and clear. 4 Reassembling your mask 1. Align and insert the frame connector tab (a) into the cushion slot (b) and press down until it clicks. Repeat on the other side. 2. Attach the elbow to the top of the frame until it clicks. 3. With the ResMed logo facing outwards and up, insert the upper headgear straps into the frame from the inside and fold the fastening tabs over to secure. Note: If the elbow ring detaches, re-insert into the top of the frame. WARNING • • • • • Magnets are used in the lower headgear straps and the frame of the mask. Ensure the headgear and frame is kept at least 2" (50 mm) away from any active medical implant (eg, pacemaker or defibrillator) to avoid possible effects from localized magnetic fields. The magnetic field strength is less than 400 mT. The mask contains safety features, the exhaust vent holes and anti-asphyxia valves, to enable normal breathing and exhaust exhaled breath. Occlusion of the exhaust vent holes or anti-asphyxia valves needs to be prevented to avoid having an adverse effect on the safety and quality of the therapy. Regularly inspect the vent holes and antiasphyxia valves to ensure they are kept clean, clear of blockages and are not damaged. Only use compatible CPAP or bi-level therapy devices or accessories. The technical specifications of the mask are provided for healthcare professionals to determine compatible devices. Use in combination with incompatible medical devices can decrease the safety or alter the performance of the mask. The mask is not suitable for patients requiring life support ventilation. Regularly clean your mask and its components to maintain the quality of your mask and to prevent the growth of germs that can adversely affect your health. English 5 WARNING • • • • • The mask must be used under qualified supervision for patients who are unable to remove the mask by themselves. The mask may not be suitable for those prone to aspiration. The mask should not be used unless the device is turned on. Once the mask is fitted, ensure the device is blowing air to reduce risk of rebreathing exhaled air. Discontinue using or replace this mask if the patient has ANY adverse reaction to the use of the mask. Consult your physician or sleep therapist. The mask is not intended to be used simultaneously with nebulizer medications that are in the air path of the mask/tube. If there is any visible deterioration of a mask component (cracking, crazing, tears etc.) the component should be discarded and replaced. CAUTION • • • • • 6 Always follow cleaning instructions and only use a mild liquid detergent. Some cleaning products may damage the mask, its parts and their function, or leave harmful residual vapours. Do not use a dish washer or washing machine to clean the mask. When fitting the mask, do not overtighten the headgear as this may lead to skin redness or sores around the mask cushion. As with all masks, some rebreathing may occur at low pressures. Using a mask may cause tooth, gum or jaw soreness or aggravate an existing dental condition. If symptoms occur, consult your physician or dentist. Do not iron the headgear as the material is heat sensitive and will be damaged. Technical specifications Mask setting options: For AirSense, AirCurve or S9: Select 'Full Face'. Compatible devices: For a full list of compatible devices for this mask, see the Mask/Device Compatibility List at ResMed.com/downloads/masks. SmartStop may not operate effectively when using this mask with some CPAP or bi-level devices. Pressure-flow curve Pressure Flow (L/min) (cm H2O) 4 22 11 38 17 48 24 59 30 67 Therapy pressure: 4 to 30 cm H20 Resistance with Anti Asphyxia Valve closed to atmosphere Drop in pressure measured (nominal) at 50 L/min: 0.2 cm H2O Drop in pressure measured (nominal) at 100 L/min: 1.0 cm H2O Resistance with Anti Asphyxia Valve open to atmosphere Inspiration at 50 L/min: 0.2 cm H2O Expiration at 50 L/min: 0.4 cm H2O Anti-Asphyxia Valve open-to-atmosphere pressure: <4 cm H2O Anti-Asphyxia Valve closed-to-atmosphere pressure: <4 cm H2O Sound: Declared dual-number noise emission values in accordance with ISO4871:1996 and ISO3744:2010. A-weighted sound power level is 25 dBA, with uncertainty of 3 dBA. Aweighted pressure level at a distance of 1 m is 18 dBA, with uncertainly of 3 dBA. Environmental conditions Operating temperature: 41°F to 104°F (5°C to 40°C) Operating humidity: 15% to 95% RH non-condensing Storage and transport temperature: -4°F to 140°F (-20°C to +60°C) Storage and transport humidity: up to 95% RH non-condensing International Commission on Non-Ionizing Radiation Protection (ICNIRP) Magnets used in this mask are within ICNIRP guidelines for general public use. Service life: The service life of the mask system is dependent on the intensity of usage, maintenance, and environmental conditions to which the mask is used or stored. As this mask system and its components are modular in nature, it is recommended that the user maintain and inspect it on a regular basis, and replace the mask system or any English 7 components if deemed necessary or according to the instructions in the ‘Cleaning your mask’ section of this guide. Accessories: Soft sleeves are available as an accessory, Storage Ensure that the mask is thoroughly clean and dry before storing it for any length of time. Store the mask in a dry place out of direct sunlight. Disposal This mask does not contain any hazardous substances and may be disposed of with your normal household refuse. Reprocessing the mask between patients Only Sleep Lab Mask (SLM) variants of the mask system are intended for multi-patient reuse. When using between patients, these masks must be reprocessed according to instructions available on ResMed.com/downloads/masks. Symbols The following symbols may appear on your product or packaging: Full face mask Device setting- Full face Small frame Standard frame Large frame Cushion size - small Cushion size - medium Cushion size - small wide Cushion size - wide Caution, consult accompanying documents. Prescription only (In the US, Federal law restricts these devices to sale by or on the order of a physician.) See symbols glossary at ResMed.com/symbols. 8 Limited warranty ResMed Ltd (hereafter ’ResMed’) warrants that your ResMed mask system (including mask frame, cushion, headgear and tubing) shall be free from defects in material and workmanship from the date of purchase for the period of 90 days or in the case of disposable masks the maximum use period. This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This limited warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product; b) repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs; and c) any damage or contamination due to cigarette, pipe, cigar or other smoke. Warranty is void on product sold, or resold, outside the region of original purchase. Warranty claims on defective product must be made by the initial consumer at the point of purchase. This warranty replaces all other expressed or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you. ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from region to region. For further information on your warranty rights, contact your local ResMed dealer or ResMed office. English 9 FRANÇAIS Merci d’avoir choisi le masque facial d’AirFit™ F30i. Ce masque dispose d’une bulle qui adhère pour une parfaite étanchéité sous le nez et autour de la bouche, et d’un entourage rigide muni d’un raccord de tuyau audessus de la tête. Pour respirer de l’air frais, ce masque est doté d’orifices de ventilation dans la bulle et le coude, ainsi que de valves dans l’entourage rigide. A B Coude Entourage rigide C Bulle D Harnais 6 Bague du coude Valve anti-asphyxie (dans le raccord de l’entourage rigide à gauche et à droite) Raccord de l’entourage rigide 1 2 3 Orifices de ventilation (bulle) Orifices de ventilation (coude) Bouton latéral 7 8 9 Clip magnétique Sangle inférieure du harnais Sangle supérieure du harnais 4 5 Usage prévu Le masque AirFit F30i est prévu pour un usage par des patients pesant plus de 30 kg (66 lb) pour qui un traitement par pression positive non invasif, comme un traitement par PPC ou à deux niveaux de pression, a été prescrit. Le masque est prévu pour un usage multiple par un seul Français 1 patient à domicile ou pour un usage multiple par plusieurs patients à l’hôpital ou en milieu médical. Contre-indications L’utilisation de masques avec des composants magnétiques est contreindiquée chez les patients présentant les pathologies ou états préexistants suivants : • dispositif hémostatique métallique implanté dans le crâne suite à un anévrisme; • présence d’éclats métalliques dans un œil ou les deux yeux. Avant d’utiliser votre masque Retirez tout l’emballage et inspectez chaque composant du masque afin de détecter toute détérioration visible. Ajustement de votre masque 2 1. Raccordez le circuit respiratoire de votre appareil au coude. 2. Appuyez sur les boutons latéraux du coude et retirez ce dernier de l’entourage rigide. Mettez le coude et le circuit respiratoire de côté pour le moment. 3. Tournez les clips magnétiques et détachez-les des raccords de l’entourage rigide. 4. Placez la bulle sous votre nez et assurez-vous qu’elle repose confortablement sur votre visage. Avec le logo ResMed du harnais dirigé vers le haut, faites passer le harnais et l’entourage rigide pardessus votre tête. 5. Ramenez les sangles inférieures du harnais sous vos oreilles, puis fixez les clips magnétiques aux raccords de l’entourage rigide. 6. Détachez les bandes de fixation des sangles supérieures du harnais et tirez dessus uniformément. Répétez l’opération avec les sangles inférieures du harnais. 7. Raccordez le coude sur le dessus de l’entourage rigide. Votre masque doit être à présent positionné comme illustré. Ajustement de votre masque • Avec l’appareil sous tension et émettant un débit d’air, ajustez la position de la bulle pour qu’elle adhère de manière confortable sous votre nez en assurant la meilleure étanchéité possible. Vérifiez que la bulle ne présente aucun pli et que le harnais n’est pas entortillé. • En cas de fuite au niveau du masque, ajustez les sangles supérieures ou inférieures du harnais. Ne les serrez pas trop : ajustez-les uniquement pour obtenir une étanchéité confortable. Français 3 Détermination de la taille du masque Servez-vous du gabarit d’ajustement pour vous aider à choisir la bonne taille de bulle. Si l’entourage rigide du masque tombe vers l’arrière ou se trouve trop près de vos oreilles, essayez un entourage rigide de taille plus petite. Si l’entourage rigide du masque tombe vers l’avant ou se trouve trop près de vos yeux, essayez un entourage rigide de taille plus grande. Démontage de votre masque pour le nettoyage Si votre masque est connecté à un appareil, débranchez le circuit respiratoire du coude du masque. 1. Détachez les bandes de fixation des sangles supérieures du harnais et retirez-les de l’entourage rigide. Laissez les clips magnétiques sur les sangles inférieures du harnais. 2. Appuyez sur les boutons latéraux du coude afin de le séparer de l’entourage rigide. 3. Tenez fermement le raccord de l’entourage rigide et soulevez-le pour le déclipser de la bulle. Répétez l’opération de l’autre côté. 4 Nettoyage de votre masque En cas de détérioration visible d’un composant du masque (fissures, fendillements, déchirures, etc.), ce composant doit être jeté et remplacé. Tous les jours/Après chaque utilisation : Bulle Chaque semaine : Harnais, entourage rigide et coude 1. Faites tremper les composants dans une eau tiède avec un détergent liquide doux. 2. Nettoyez les composants à la main à l’aide d’une brosse à soies souples. Veillez en particulier à bien nettoyer les orifices de ventilation situés à la fois dans la bulle et dans le coude. 3. Rincez soigneusement les composants sous l’eau courante. 4. Laissez les composants sécher à l’abri de la lumière directe du soleil. Si les composants du masque semblent toujours sales, répétez les différentes étapes de nettoyage. Vérifiez que les orifices de ventilation et les valves anti-asphyxie sont propres et nets. Remontage de votre masque 1. Alignez et insérez la languette du raccord de l’entourage rigide (a) dans la fente de la bulle (b) et appuyez dessus jusqu’à ce qu’il s’enclenche. Répétez l’opération de l’autre côté. 2. Fixez le coude sur le dessus de l’entourage rigide jusqu’à ce qu’il s’enclenche. 3. Avec le logo ResMed dirigé vers l’extérieur et le haut, insérez les sangles supérieures du harnais dans l’entourage rigide, de l’intérieur vers l’extérieur, puis repliez les bandes de fixation pour les fermer. Remarque : Si la bague du coude se détache, insérez-la de nouveau dans le dessus de l’entourage rigide. Français 5 AVERTISSEMENT • • • • • • • • • • 6 Des aimants sont utilisés dans les sangles inférieures du harnais et dans l’entourage rigide du masque. Vérifiez que le harnais et l’entourage rigide sont maintenus à une distance minimale de 50 mm (2") de tout implant médical actif (par ex. un pacemaker ou un défibrillateur) afin d’éviter les effets possibles des champs magnétiques localisés. La force du champ magnétique est inférieure à 400 mT. Le masque est doté de dispositifs de sécurité – les orifices de ventilation et les valves anti-asphyxie – afin de permettre à la personne de respirer normalement et d’évacuer l’air expiré. L’obstruction des orifices de ventilation ou des valves anti-asphyxie doit être à tout prix évitée pour ne pas affecter la sécurité et la qualité du traitement. Inspectez régulièrement les orifices de ventilation et les valves anti-asphyxie pour veiller à leur propreté, leur bon état et leur dégagement. N’utilisez que des accessoires ou des appareils de traitement par PPC ou à deux niveaux de pression compatibles. Les caractéristiques techniques du masque accompagnent ce dernier pour que les professionnels de santé puissent déterminer les appareils compatibles. Utiliser le masque avec des appareils médicaux incompatibles peut compromettre la sécurité ou altérer son efficacité. Le masque ne convient pas aux patients nécessitant une ventilation support de vie. Nettoyez régulièrement votre masque et ses composants pour maintenir leur qualité et éviter la prolifération de germes potentiellement nuisibles à votre santé. Le masque doit être utilisé sous le contrôle d’une personne qualifiée lorsque le patient n’est pas en mesure de l’enlever de lui-même. Le masque peut ne pas convenir aux patients sujets aux aspirations. Le masque ne doit être porté que si l’appareil est sous tension. Une fois le masque en place, assurez-vous que l’appareil émet un débit d’air afin de réduire le risque de réinhalation d’air expiré. En cas de réaction indésirable QUELCONQUE au masque de la part du patient, cessez de l’utiliser ou remplacez-le. Consultez votre médecin ou un spécialiste du sommeil. Ce masque n’est pas destiné à être utilisé conjointement à des médicaments normalement administrés par nébuliseur qui se trouvent dans tout composant du passage de l’air du masque ou du tuyau. En cas de détérioration visible d’un composant du masque (fissures, fendillements, déchirures, etc.), ce composant doit être jeté et remplacé. ATTENTION • • • • • Veillez à toujours respecter les instructions de nettoyage et à n’utiliser qu’un détergent doux. Certains produits de nettoyage risquent d’endommager le masque, ses pièces et leur fonctionnement ou de laisser des vapeurs résiduelles nocives. Ne nettoyez pas le masque à la machine à laver ou au lave-vaisselle. Lors de l’ajustement du masque, veillez à ne pas trop serrer le harnais; cela peut mener à l’apparition de rougeurs et d’irritations cutanées autour de la bulle du masque. Comme c’est le cas avec tous les masques, une réinhalation peut se produire à une pression basse. L’utilisation d’un masque peut provoquer des douleurs dans les dents, les gencives ou la mâchoire ou aggraver un problème dentaire existant. Si ces symptômes apparaissent, consultez un médecin ou un dentiste. Ne repassez pas le harnais au risque d’endommager le matériau avec lequel il est fabriqué, celui-ci étant sensible à la chaleur. Caractéristiques techniques Fonction de sélection du type de masque : Pour AirSense, AirCurve ou S9 : Sélectionnez « Facial ». Appareils compatibles : Pour une liste complète des appareils compatibles avec ce masque, veuillez consulter la liste de compatibilité appareil/masque sur le site www.ResMed.com/downloads/masks. La fonction SmartStop peut ne pas fonctionner correctement lorsque ce masque est utilisé avec certains appareils de PPC ou d'aide inspiratoire avec PEP. Courbe pression/débit : Pression Débit (cm H2O) (l/min) 4 22 11 38 17 48 24 59 30 67 Pression de traitement : 4 à 30 cm H2O Résistance avec la valve anti-asphyxie fermée Chute de pression mesurée (nominale) à 50 l/min : 0,2 cm H2O Chute de pression mesurée (nominale) à 100 l/min : 1,0 cm H2O Français 7 Résistance avec la valve anti-asphyxie ouverte Inspiration à 50 l/min : 0,2 cm H2O Expiration à 50 l/min : 0,4 cm H2O Pression avec la valve anti-asphyxie ouverte : ≤ 4 cm H2O Pression avec la valve anti-asphyxie fermée : ≤ 4 cm H2O Niveau sonore : Valeurs d’émissions sonores à deux chiffres déclarées conformément aux normes ISO4871:1996 et ISO3744:2010. Le niveau de puissance acoustique pondéré A est de 25 dBA avec une incertitude de 3 dBA. Le niveau de pression acoustique pondéré A mesuré à une distance d’1 m est de 18 dBA avec une incertitude de 3 dBA. Conditions ambiantes Température de fonctionnement : de 41°F à 104°F (de 5°C à 40°C) Humidité de fonctionnement : de 15 % à 95 % HR sans condensation Température de stockage et de transport : de -4°F à 140°F (de -20°C à +60°C) Humidité de stockage et de transport : jusqu’à 95 % HR sans condensation Commission internationale de protection contre les rayonnements non ionisants (CIPRNI) Les éléments magnétiques de ce masque respectent les directives de la CIPRNI pour une utilisation grand public. Durée de vie : La durée de vie du masque dépend de son utilisation, de l’entretien et des conditions ambiantes dans lesquelles le masque est utilisé et rangé. Compte tenu de la nature modulaire du masque et de ses composants, il est recommandé à l’utilisateur d’entretenir et d’inspecter régulièrement le masque et de le remplacer ou de remplacer tout composant le cas échéant ou conformément aux instructions indiquées dans la section « Nettoyage de votre masque » de ce guide. Accessoires : Des gaines souples sont proposées comme accessoires. Stockage Veillez à ce que le masque soit complètement propre et sec avant de le ranger. Rangez le masque dans un endroit sec à l’abri de la lumière directe du soleil. Élimination Ce masque ne contient aucune substance dangereuse et peut être jeté avec les ordures ménagères. Retraitement du masque entre les patients Seules les versions SLM (Sleep Lab Mask) du masque sont destinées à un usage multiple par plusieurs patients. Avant d’être utilisés par un autre patient, ces masques doivent être retraités conformément aux consignes disponibles sur ResMed.com/downloads/masks. 8 Symboles Les symboles suivants peuvent figurer sur votre produit ou sur son emballage : Masque facial Réglage de l’appareil - Facial Entourage rigide de petite taille Entourage rigide de taille standard Entourage rigide de grande taille Taille de la bulle - petite Taille de la bulle - moyenne Taille de la bulle - petite-large Taille de la bulle - large N'est pas fabriqué avec du latex de caoutchouc naturel Attention, veuillez consulter la documentation jointe. Sur ordonnance uniquement (selon la loi fédérale américaine, cet appareil ne peut être vendu aux États-Unis que par un médecin ou sur prescription médicale). Reportez-vous au glossaire des symboles au ResMed.com/symbols. Garantie limitée ResMed Ltd (ci-après « ResMed ») garantit votre masque ResMed (y compris l’entourage rigide, la bulle, le harnais et le circuit respiratoire) contre tout défaut de matériaux et de main-d’œuvre pour une période de 90 jours à compter de la date d’achat ou, dans le cas d’un masque jetable, pour la période d’utilisation maximale. Seul le client initial est couvert par la présente garantie. Celle-ci n’est pas cessible. En cas de défaillance du produit dans des conditions normales d’utilisation, ResMed, à son entière discrétion, répare ou remplace le produit défectueux ou toute pièce. La présente garantie limitée ne couvre pas : a) tout dommage résultant d’une utilisation incorrecte, d’un usage abusif ou d’une modification ou transformation opérée sur le produit; b) les réparations effectuées par tout service de réparation sans l’autorisation expresse de ResMed; et c) tout dommage ou contamination causé par de la fumée de cigarette, de pipe, de cigare ou autre. La garantie est annulée pour les produits vendus ou revendus dans un pays autre que celui où ils ont été achetés à l’origine. Français 9 Les réclamations au titre de la garantie pour les produits défectueux doivent être présentées au lieu d’achat par le client initial. La présente garantie remplace toute autre garantie expresse ou implicite, y compris toute garantie implicite de qualité marchande ou d’adéquation à un usage particulier. Certains pays ou États n’autorisent pas les limitations de durée pour les garanties implicites; il est donc possible que la limitation susmentionnée ne s’applique pas à votre cas. La société ResMed ne peut être tenue responsable de tout dommage accessoire ou indirect résultant de la vente, de l’installation ou de l’utilisation de tout produit ResMed. Certains pays ou États n’autorisent ni l’exclusion ni la limitation des dommages accessoires ou indirects; il est donc possible que la limitation susmentionnée ne s’applique pas à votre cas. La présente garantie vous octroie des droits reconnus par la loi. Vous pouvez également détenir d’autres droits qui varient en fonction du pays où vous habitez. Pour de plus amples informations sur vos droits de garantie, veuillez contacter ResMed ou votre revendeur local ResMed. 10 ESPAÑOL Gracias por elegir la mascarilla facial AirFit™ F30i. En esta mascarilla se utiliza una almohadilla que permite sellar debajo de la nariz y alrededor de la boca, y un armazón que se conecta con el tubo por encima de la cabeza. Para que se respire aire puro, esta mascarilla tiene orificios de ventilación en la almohadilla y el codo, así como válvulas en el armazón. A B C Codo Armazón Almohadilla 4 5 D 1 Arnés Orificios de ventilación (almohadilla) Orificios de ventilación (codo) Botón lateral 6 7 Aro del codo Válvula antiasfixia (dentro del lado izquierdo y derecho de la conexión del armazón) Conexión del armazón Broche magnético 8 9 Correa inferior del arnés Correa superior del arnés 2 3 Uso indicado La mascarilla AirFit F30i está diseñada para ser utilizada en pacientes que pesen más de 30 kg (66 lb) a los que se les haya indicado tratamiento no invasivo con presión positiva en las vías respiratorias (PAP), como un tratamiento CPAP o binivel. La mascarilla está diseñada para ser utilizada en repetidas ocasiones por un único paciente en el domicilio, y por varios pacientes en hospitales o instituciones. Español 1 Contraindicaciones El uso de las mascarillas con componentes magnéticos está contraindicado en pacientes que tengan las siguientes afecciones preexistentes: • una pinza hemostática de metal implantada en la cabeza para reparar un aneurisma; • esquirlas metálicas en uno o ambos ojos. Antes de utilizar la mascarilla Quite todo el embalaje e inspeccione cada una de las piezas para asegurarse de que no haya signos visibles de deterioro. Colocación de la mascarilla 2 1. Conecte el tubo de aire del equipo al codo. 2. Apriete los botones laterales del codo y tire para separarlo del armazón. Deje el codo y el tubo de aire a un lado por ahora. 3. Gire ambos broches magnéticos y tire de ellos para separarlos de las conexiones del armazón. 4. Colóquese la almohadilla debajo de la nariz y asegúrese de que quede cómodamente apoyada sobre el rostro. Coloque el arnés con el logotipo de ResMed mirando hacia arriba y pase el arnés y el armazón por encima de la cabeza. 5. Pase las correas inferiores del arnés por debajo de las orejas y conecte los broches magnéticos a las conexiones del armazón. 6. Despegue las lengüetas de sujeción de las correas superiores del arnés y tire de forma pareja. Repita ese paso con las correas inferiores del arnés. 7. Conecte el codo a la parte superior del armazón. La mascarilla debe quedar colocada como se muestra en la figura. Ajuste de la mascarilla • Con el equipo encendido y suministrando aire, corrija la posición de la almohadilla para lograr el sellado que le resulte más cómodo debajo de la nariz. Asegúrese de que la almohadilla no esté plegada y el arnés no esté torcido. • Para resolver las fugas que pueda haber por la mascarilla, ajuste las correas superiores o inferiores del arnés. Ajústelas lo que sea necesario para obtener un sellado cómodo: no las apriete demasiado. Español 3 Selección del tamaño de la mascarilla Use el modelo de ajuste para seleccionar una almohadilla que tenga el tamaño correcto. Si el armazón de la mascarilla se le desliza hacia atrás por la cabeza o le queda apoyado muy cerca de las orejas, pruebe con uno de tamaño más pequeño. Si la mascarilla se le desliza hacia adelante por la cabeza o le queda apoyada muy cerca de los ojos, pruebe con un armazón más grande. Desmontaje de la mascarilla para limpiarla Si la mascarilla está conectada a un equipo, tome el tubo de aire de este último y desconéctelo del codo de la mascarilla. 1. Despegue las lengüetas de sujeción de las correas superiores del arnés y tire de ellas para separarlas del armazón. Mantenga los broches magnéticos unidos a las correas inferiores del arnés. 2. Apriete los botones laterales ubicados en el codo y desconéctelo del armazón. 3. Tome la conexión del armazón y levántela para desabrocharla de la almohadilla. Haga lo mismo en el otro lado. 4 Limpieza de la mascarilla Si advierte algún signo de deterioro en alguna de las piezas de la mascarilla (como grietas, rajaduras, roturas, etc.), la pieza debe desecharse y sustituirse por otra. Diariamente/después de cada uso: almohadilla Semanal: arnés, armazón y codo. 1. Sumerja las piezas en agua tibia con detergente líquido suave. 2. Lave las piezas a mano con un cepillo de cerdas suaves. Preste especial atención a los orificios de ventilación tanto de la almohadilla como del codo. 3. Enjuague las piezas por completo bajo agua de grifo. 4. Deje secar las piezas al aire, sin exponerlas a la luz solar directa. Si las piezas de la mascarilla no se ven limpias, repita los pasos de limpieza. Compruebe que los orificios de ventilación y las válvulas antiasfixia estén limpios y despejados. Montaje de la mascarilla 1. Coloque la lengüeta de la conexión del armazón (a) a la misma altura que la ranura de la almohadilla (b), insértela en ella y presione hacia abajo hasta que haga clic. Haga lo mismo en el otro lado. 2. Conecte el codo en la parte superior del armazón hasta que haga clic. 3. Con el logotipo de ResMed mirando hacia afuera y hacia arriba, inserte las correas superiores del arnés en el armazón desde el lado interior y doble las lengüetas de sujeción para sujetar. Nota: Si el aro del codo se desconecta, vuelva a insertarlo en la parte superior del armazón. Español 5 ADVERTENCIA • • • • • • • • • • 6 Se utilizan imanes en las correas inferiores del arnés y en el armazón de la mascarilla. Asegúrese de que el arnés y el armazón se mantengan por lo menos a 50 mm (2 pulg.) de distancia de cualquier implante médico activo (p. ej. un marcapaso o desfibrilador), para evitar posibles efectos provocados por los campos magnéticos localizados. La intensidad de los campos magnéticos es inferior a 400 mT. La mascarilla tiene características de seguridad —los orificios de ventilación y las válvulas antiasfixia—, para que se pueda respirar normalmente y expulsar el aire exhalado. Es necesario evitar que los orificios de ventilación o las válvulas antiasfixia se obstruyan, para no perjudicar la seguridad y la calidad del tratamiento. Inspeccione con regularidad los orificios de ventilación y las válvulas antiasfixia para cerciorarse de que estén limpios, no tengan obstrucciones y no estén dañados. Utilice solo accesorios o equipos de tratamiento binivel o CPAP que sean compatibles. Las especificaciones técnicas de la mascarilla se proporcionan para que los profesionales de salud puedan determinar qué equipos son compatibles. Usarla con equipos médicos incompatibles puede disminuir la seguridad o alterar el rendimiento de la mascarilla. La mascarilla no es adecuada para utilizarla en pacientes que necesitan ventilación de apoyo vital. Limpie la mascarilla y sus piezas con regularidad para preservar la calidad de la mascarilla y evitar la presencia de microbios que pueden ser perjudiciales para la salud. La mascarilla debe ser utilizada bajo supervisión calificada cuando los pacientes no sean capaces de retirarla por sí mismos. Es posible que la mascarilla no sea apropiada para quienes sean propensos a la aspiración. La mascarilla no debe utilizarse a menos que el equipo se encuentre encendido. Una vez que la mascarilla esté colocada, compruebe que el equipo esté suministrando aire para reducir el riesgo de volver a inspirar el aire exhalado. Deje de usar esta mascarilla o reemplácela si el paciente presenta CUALQUIER reacción adversa a su utilización. Consulte al médico o al especialista en medicina del sueño. No se debe utilizar la mascarilla junto con medicamentos para nebulizadores ubicados en el circuito de aire de la mascarilla o el tubo. Si advierte algún signo de deterioro en alguna de las piezas de la mascarilla (como grietas, rajaduras, roturas, etc.), la pieza debe desecharse y sustituirse por otra. PRECAUCIÓN • • • • • Siga siempre las instrucciones de limpieza y utilice únicamente un detergente líquido suave. Algunos productos de limpieza pueden dañar la mascarilla, sus piezas y su funcionamiento, o dejar vapores residuales dañinos. No lave la mascarilla en un lavavajillas o lavarropas. Cuando coloque la mascarilla, no apriete demasiado el arnés, ya que esto puede provocar enrojecimiento de la piel o llagas alrededor de la almohadilla. Como ocurre con todas las mascarillas, si la presión es baja se puede volver a inhalar parte del aire exhalado. Utilizar una mascarilla puede provocar dolor en los dientes o muelas, en las encías o en la mandíbula, o agravar una afección dental ya existente. Si se presentan síntomas, consulte al médico o al odontólogo. No planche el arnés, ya que está hecho de un material sensible al calor y se dañará. Especificaciones técnicas Opciones de configuración de la mascarilla: en el AirSense, AirCurve o S9, seleccione Facial. Equipos compatibles: para obtener una lista completa de los equipos compatibles con esta mascarilla, consulte la lista de compatibilidad entre mascarillas y equipos en ResMed.com/downloads/masks. Es posible que SmartStop no funcione eficazmente al usar esta mascarilla con algunos equipos CPAP o binivel. Curva de flujo en función de la presión Presión Flujo (cm H2O) (l/min) 4 22 11 38 17 48 24 59 30 67 Presión de tratamiento: de 4 a 30 cm H2O Resistencia con válvula antiasfixia cerrada a la atmósfera Caída de presión medida (nominal) a 50 l/min: 0,2 cm H2O Caída de presión medida (nominal) a 100 l/min: 1,0 cm H2O Resistencia con válvula antiasfixia abierta a la atmósfera Inspiración a 50 l/min: 0,2 cm H2O Espiración a 50 l/min: 0,4 cm H2O Español 7 Presión con válvula antiasfixia abierta a la atmósfera: <4 cm H2O Presión con válvula antiasfixia cerrada a la atmósfera: <4 cm H2O Sonido: valores de emisión sonora declarados mediante dos números de conformidad con la ISO4871:1996 y la ISO3744:2010. El nivel de potencia acústica ponderado según la escala A es de 25 dBA, con una incertidumbre de 3 dBA. El nivel de presión acústica ponderado según la escala A a una distancia de 1 m es de 18 dBA, con una incertidumbre de 3 dBA. Condiciones ambientales Temperatura de funcionamiento: de 5 °C a 40 °C (de 41 °F a 104 °F) Humedad de funcionamiento: de 15 % a 95 % de HR sin condensación Temperatura de almacenamiento y transporte: de -20 °C a +60 °C (de -4 °F a 140 °F) Humedad de almacenamiento y transporte: hasta 95 % de HR sin condensación Comisión Internacional de Protección contra la Radiación No Ionizante (ICNIRP) Los imanes que se usan en esta mascarilla cumplen con las directrices de la ICNIRP para uso por parte del público en general. Vida útil: la vida útil del sistema de la mascarilla depende de la intensidad con que se lo use, el mantenimiento que se le dé y las condiciones ambientales en que la mascarilla se use o almacene. Dado que este sistema de mascarilla y sus piezas son de naturaleza modular, se recomienda al usuario que le dé mantenimiento y lo inspeccione con regularidad, y que cambie el sistema o cualquiera de sus piezas si lo considera necesario o si así se indica en las instrucciones que se brindan en la sección “Limpieza de la mascarilla” del presente manual. Accesorios: hay fundas blandas disponibles como accesorio. Almacenamiento Asegúrese de que la mascarilla esté completamente limpia y seca antes de guardarla durante cualquier período considerable de tiempo. Guarde la mascarilla en un lugar seco, fuera de la luz solar directa. Eliminación La mascarilla no contiene ninguna sustancia peligrosa y puede eliminarse con los residuos domésticos habituales. Reprocesamiento de la mascarilla de un paciente a otro Solo las variantes SLM (para usar en laboratorios del sueño) del sistema de la mascarilla están diseñadas para utilizarse en más de un paciente. Cuando se vaya a cambiar de paciente, estas mascarillas se deben reprocesar de conformidad con las instrucciones que se brindan en ResMed.com/downloads/masks. 8 Símbolos Los símbolos que figuran a continuación pueden aparecer en el producto o en el envoltorio: Mascarilla facial Configuración del equipo: facial Armazón pequeño Armazón estándar Armazón grande Tamaño de la almohadilla: pequeño Tamaño de la almohadilla: mediano Tamaño de la almohadilla: ancho y pequeño Tamaño de la almohadilla: ancho No hecho con látex de caucho natural Precaución: consulte los documentos adjuntos. Solo con receta (en EE. UU., la ley federal exige que estos equipos sean vendidos únicamente por un médico o por su orden) Véase el glosario de símbolos en ResMed.com/symbols. Garantía limitada ResMed Ltd (en adelante “ResMed”) garantiza que su sistema de mascarilla ResMed (incluidos el armazón, la almohadilla, el arnés y el tubo de la mascarilla) estará libre de todo defecto de material y mano de obra a partir de la fecha de adquisición y durante un período de 90 días o, en el caso de las mascarillas desechables, durante el período máximo de uso. Esta garantía solo tiene validez para el consumidor inicial y no es transferible. Si el producto falla en condiciones de utilización normales, ResMed reparará o reemplazará, a opción de ResMed, el producto defectuoso o cualquiera de sus piezas. Esta garantía limitada no cubre: a) ningún daño provocado por la utilización indebida, el abuso, la modificación o la alteración del producto; b) reparaciones llevadas a cabo por cualquier organización dedicada a la reparación que no haya sido expresamente autorizada por ResMed para hacer dichas reparaciones, y c) ningún daño o contaminación provocados por humo de cigarrillo, pipa, habano u otros tipos de humo. La garantía queda anulada si el producto se vende o revende fuera de la región de compra original. Español 9 Las reclamaciones de garantía con respecto a productos defectuosos deben ser realizadas por el consumidor original en el punto de compra. Esta garantía reemplaza cualquier otra garantía expresa o implícita, incluida cualquier garantía implícita de comerciabilidad o idoneidad para un propósito en particular. Algunas regiones o estados no permiten limitaciones respecto de la duración de una garantía implícita, por lo que es posible que la limitación estipulada anteriormente no sea aplicable en su caso particular. ResMed no se responsabilizará de ningún daño incidental ni emergente que se diga que ocurrió a consecuencia de la venta, la instalación o el uso de alguno de los productos de ResMed. Algunas regiones o estados no permiten la exclusión o la limitación de los daños incidentales o emergentes, por lo que es posible que la limitación estipulada anteriormente no sea aplicable en su caso particular. La presente garantía le otorga derechos legales específicos, y es posible que usted también tenga otros derechos que pueden variar de una región a otra. Para obtener más información acerca de los derechos de garantía, póngase en contacto con el distribuidor de ResMed o la sucursal de ResMed de su zona. 10 PORTUGUÊS Obrigado por escolher a máscara oronasal AirFit™ F30i. Essa máscara possui uma almofada que veda a região abaixo do nariz e ao redor da boca, além de uma armação com uma conexão de tubo acima da cabeça. Para respirar ar fresco, a máscara conta com orifícios do respiradouro na almofada e no cotovelo, e válvulas na armação. A B C Cotovelo Armação Almofada 4 5 D 1 Arnês Orifícios do respiradouro (almofada) Orifícios do respiradouro (cotovelo) 6 7 Anel do cotovelo Válvula antiasfixia (dentro da parte esquerda e direita do conector da armação) Conector da armação Presilha magnética 8 Correia inferior do arnês Botão lateral 9 Correia superior do arnês 2 3 Uso previsto A máscara AirFit F30i foi projetada para pacientes com peso superior a 30 kg (66 lb) que receberam prescrição de terapia de pressão positiva não invasiva nas vias respiratórias (PAP), como CPAP ou terapia de dois níveis. A máscara foi projetada para reutilização por um único paciente em Português 1 ambiente doméstico e reutilização por vários pacientes em ambiente hospitalar/institucional. Contraindicações O uso de máscaras com componentes magnéticos é contraindicado em pacientes que já apresentem as seguintes condições: • clipes hemostáticos metálicos implantados na cabeça para reparação de um aneurisma • fragmentos metálicos em um ou em ambos os olhos. Antes de usar a máscara Remova todas as embalagens e verifique se algum componente da máscara apresenta deterioração visível. Colocação da máscara 2 1. Conecte o tubo de ar do dispositivo ao cotovelo. 2. Pressione os botões laterais no cotovelo e puxe a peça da armação. Deixe o cotovelo e o tubo de ar de lado agora. 3. Gire e puxe as duas presilhas magnéticas para fora dos conectores da armação. 4. Coloque a almofada sob o nariz e veja se está confortável em seu rosto. Com o logotipo da ResMed no arnês voltado para cima, passe o arnês e a armação sobre a cabeça. 5. Passe as correias inferiores do arnês por baixo das orelhas e prenda as presilhas magnéticas aos conectores da armação. 6. Solte as pontas de fixação das correias superiores do arnês e puxe uniformemente. Repita com as correias inferiores do arnês. 7. Prenda o cotovelo à parte de cima da armação. A máscara deve ser posicionada conforme mostrado. Ajuste da máscara • Com o dispositivo ligado e soprando ar, ajuste a posição da almofada para que vede do modo mais confortável possível sob o nariz. Certifique-se de que a almofada não esteja dobrada e o arnês não esteja torcido. • Para consertar fugas da máscara, ajuste as correias inferiores e superiores do arnês. Ajuste apenas o suficiente para uma vedação confortável e não aperte demais. Tamanho da máscara Use o modelo de ajuste para auxiliar na seleção da almofada no tamanho correto. Se a armação da máscara ficar muito para trás da cabeça ou muito próxima dos ouvidos, experimente um tamanho de armação menor. Se a Português 3 máscara ficar muito para a frente da cabeça ou muito próxima dos olhos, experimente um tamanho de armação maior. Desmontagem da máscara para limpeza Se a máscara estiver conectada a um dispositivo, desconecte o tubo de ar do dispositivo pelo cotovelo. 1. Solte as pontas de fixação nas correias superiores do arnês e puxe para liberá-las da armação. Mantenha as presilhas magnéticas fixadas nas correias inferiores do arnês. 2. Pressione os botões laterais no cotovelo e solte a peça da armação. 3. Aperte o conector da armação e levante para soltá-lo da almofada. Repita do outro lado. 4 Limpeza da máscara Se for observado que algum componente da máscara se encontra deteriorado (quebrado, rachado, roto, etc.), tal componente deve ser descartado e substituído. Diariamente/Após cada utilização: Almofada Semanalmente: Arnês, armação e cotovelo 1. Mergulhe os componentes em água morna com um detergente líquido suave. 2. Lave manualmente os componentes com uma escova de cerdas macias. Preste muita atenção aos orifícios do respiradouro na almofada e no cotovelo. 3. Enxágue completamente os componentes com água corrente. 4. Deixe os componentes secar sem contato com luz solar direta. Se os componentes da máscara não estiverem visivelmente limpos, repita as etapas de limpeza. Certifique-se de que os orifícios do respiradouro e as válvulas antiasfixia estejam limpos e desobstruídos. Remontagem da máscara 1. Alinhe e insira a aba do conector da armação (a) à fenda da almofada (b) e pressione até ouvir um clique. Repita do outro lado. 2. Prenda o cotovelo à parte de cima da armação até ouvir um clique. 3. Com o logotipo da ResMed voltado para cima e para fora, insira as correias superiores do arnês na armação de dentro para fora e dobre as pontas de fixação para prender. Nota: Se o anel do cotovelo se soltar, insira novamente na parte de cima da armação. Português 5 AVISO • • • • • • • • • • 6 São utilizados ímãs nas correias inferiores do arnês e na armação da máscara. Mantenha o arnês e a armação a no mínimo 50 mm (2 pol.) de distância de qualquer implante médico ativo (por exemplo, marca-passo ou desfibrilador) para evitar os possíveis efeitos de campos magnéticos localizados. A intensidade do campo magnético é inferior a 400 mT. A máscara contém recursos de segurança, os orifícios do respiradouro e as válvulas antiasfixia, para permitir que o paciente respire e exale o ar normalmente. Evite a oclusão dos orifícios do respiradouro ou das válvulas antiasfixia para impedir a ocorrência de efeitos adversos na segurança e qualidade do tratamento. Inspecione frequentemente os orifícios do respiradouro e as válvulas antiasfixia para garantir que eles estejam limpos, sem qualquer bloqueio e não estejam danificados. Utilize apenas dispositivos e acessórios terapêuticos CPAP ou de dois níveis compatíveis. As especificações técnicas da máscara são fornecidas para profissionais de saúde para determinar dispositivos compatíveis. O uso em combinação com dispositivos médicos incompatíveis pode diminuir a segurança ou alterar o desempenho da máscara. A máscara não é adequada para pacientes que necessitam de ventilação de suporte à vida. Limpe regularmente sua máscara e seus componentes para manter a qualidade de sua máscara e para impedir a proliferação de germes que podem afetar negativamente sua saúde. Pacientes que não consigam retirar a máscara sozinhos devem usar a máscara sob a supervisão de pessoas qualificadas. A máscara pode não ser adequada para pessoas predispostas a aspiração. A máscara só deve ser usada se o dispositivo estiver ligado. Depois de colocada a máscara, certifique-se de que o dispositivo esteja soprando ar, para reduzir o risco de reinalação do ar exalado. Interrompa o uso ou substitua essa máscara, se o paciente tiver QUALQUER reação adversa ao uso da máscara. Consulte seu médico ou terapeuta de sono. A máscara não deve ser usada concomitantemente com medicamentos de nebulizador situados na passagem de ar da máscara/do tubo. Se for observado que algum componente da máscara se encontra deteriorado (quebrado, rachado, roto, etc.), tal componente deve ser descartado e substituído. PRECAUÇÃO • • • • • Siga sempre as instruções de limpeza e use apenas detergente líquido suave. Alguns produtos de limpeza podem danificar a máscara, suas peças e funções, ou deixar vapores residuais nocivos. Não use máquinas de lavar louça ou roupa para limpar a máscara. Ao colocar a máscara, não aperte demais o arnês, pois isso pode causar vermelhidão ou feridas na pele ao redor da almofada da máscara. Como em todas as máscaras, poderá ocorrer uma certa reinalação em pressões baixas. O uso da máscara pode deixar os dentes, a gengiva ou a mandíbula doloridos, ou agravar algum problema odontológico existente. Se aparecerem sintomas, consulte seu médico ou dentista. Não passe o arnês a ferro, pois o material é sensível ao calor e será danificado. Especificações técnicas Opções de configuração da máscara: Para AirSense, AirCurve ou S9: Selecione "Facial". Dispositivos compatíveis: Para obter uma lista completa de dispositivos compatíveis com esta máscara, consulte a Lista de Compatibilidade entre Máscara/Dispositivo em ResMed.com/downloads/masks. O SmartStop poderá não funcionar efetivamente quando esta máscara for usada com certos dispositivos CPAP ou de dois níveis. Curva de fluxo/pressão Pressão Fluxo (cm H2O) (l/min) 4 22 11 38 17 48 24 59 30 67 Pressão terapêutica: 4 a 30 cm H20 Resistência com válvula antiasfixia fechada para o ambiente Medição da queda de pressão (nominal) a 50 l/min: 0,2 cm H2O Medição da queda de pressão (nominal) a 100 l/min: 1,0 cm H2O Resistência com válvula antiasfixia aberta para o ambiente Inspiração a 50 l/min: 0,2 cm H2O Português 7 Expiração a 50 l/min: 0,4 cm H2O Pressão com válvula antiasfixia aberta para o ambiente: <4 cm H2O Pressão com válvula antiasfixia fechada para o ambiente: <4 cm H2O Som: Valores declarados de emissão de ruído expressos por um número duplo em conformidade com a ISO4871:1996 e ISO3744:2010. O nível de potência sonora ponderado em A é de 25 dBA, com imprecisão de 3 dBA. O nível de pressão sonora ponderado em A a uma distância de 1 m é de 18 dBA, com imprecisão de 3 dBA. Condições ambientais Temperatura operacional: 5°C a 40°C (41°F a 104°F) Umidade operacional: 15% a 95% UR sem condensação Temperatura de armazenamento e transporte: -20°C a +60°C (-4°F a 140°F) Umidade de armazenamento e transporte: até 95% UR sem condensação Comissão Internacional de Proteção Contra Radiação Não Ionizante (ICNIRP) Os ímãs usados nesta máscara obedecem às diretrizes da ICNIRP para uso pelo público em geral. Vida útil: A vida útil do sistema de máscara depende da intensidade de uso, da manutenção e das condições ambientais nas quais a máscara é utilizada ou armazenada. Como o sistema de máscara e seus componentes têm natureza modular, é recomendável que o usuário os conserve e examine regularmente e substitua o sistema ou qualquer componente em caso de necessidade ou de acordo as instruções na seção “Limpeza da máscara” deste guia. Acessórios: Luvas maleáveis estão disponíveis como acessórios. Armazenamento Certifique-se de que a máscara esteja totalmente limpa e seca antes de guardá-la por qualquer período de tempo. Guarde a máscara em local seco, ao abrigo da luz solar direta. Eliminação Esta máscara não contém qualquer substância perigosa e pode ser eliminada com o lixo doméstico. Reprocessamento da máscara entre pacientes Somente as variantes SLM (Sleep Lab Mask) do sistema de máscara foram projetadas para reutilização por vários pacientes. Em caso de uso em outros pacientes, estas máscaras têm de ser reprocessadas de acordo com as instruções disponíveis em ResMed.com/downloads/masks. 8 Símbolos Os símbolos a seguir podem aparecer no seu produto ou na embalagem: Configuração do dispositivo Máscara oronasal oronasal Armação pequena Armação padrão Armação grande Tamanho da almofada - pequena Tamanho da almofada - média Tamanho da almofada - pequena larga Tamanho da almofada - larga Não é feito com látex de borracha natural Precaução, consulte os documentos anexos. Somente com prescrição (nos EUA, a lei federal limita a venda destes dispositivos a médicos ou por prescrição médica). Consulte o glossário de símbolos em ResMed.com/symbols. Garantia limitada A ResMed Ltd (daqui por diante, "ResMed") garante que seu sistema de máscara ResMed (incluindo a armação da máscara, a almofada, o arnês e os tubos) está livre de defeitos de material e mão de obra por um período de 90 dias a partir da data de compra ou, no caso de máscaras descartáveis, pelo período máximo de uso. Esta garantia só está disponível para o consumidor inicial. Não é transferível. Se o produto apresentar defeito em condições normais de utilização, a ResMed efetuará, por opção própria, o reparo ou a substituição do produto defeituoso ou de qualquer um de seus componentes. Esta garantia limitada não cobre: a) danos provocados por uso incorreto, uso abusivo, modificação ou alteração do produto; b) consertos efetuados por empresas de assistência técnica que não tenham sido expressamente autorizadas pela ResMed para efetuar tais serviços; e c) danos ou contaminações provocadas por fumaça de cigarro, cachimbo, charuto ou outras fontes. A garantia deixa de ser válida se o produto for vendido, ou revendido, fora da região de compra original. Português 9 Os pedidos de reparo ou substituição de um produto defeituoso no âmbito da garantia devem ser feitos pelo consumidor original no local de compra. Esta garantia substitui todas as outras, explícitas ou implícitas, incluindo qualquer garantia implícita de comerciabilidade ou de adequabilidade para determinado fim. Algumas regiões ou estados não permitem limitações de tempo sobre a duração de uma garantia implícita, pelo que a limitação acima pode não se aplicar ao seu caso. A ResMed não será responsabilizada por quaisquer danos incidentais ou consequentes reivindicados como decorrentes da venda, instalação ou uso de qualquer produto ResMed. Algumas regiões ou estados não permitem a exclusão ou limitação de danos incidentais ou consequentes; assim, a limitação acima pode não se aplicar ao seu caso. Esta garantia confere-lhe direitos legais específicos, e você pode ter outros direitos que variam de região para região. Para obter mais informações sobre seus direitos de garantia, entre em contato com o revendedor ou escritório local da ResMed. 10 en Full face mask fr Masque facial es Mascarilla facial Device setting pt Máscara oronasal Full face ResMed Pty Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia See ResMed.com for other ResMed locations worldwide. AirFit, AirSense, AirCurve and S9 are trademarks and/or registered trademarks of the ResMed family of companies. For patent and other intellectual property information, see ResMed.com/ip. © 2020 ResMed. 638225/1 2020-02 ResMed.com 638225
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41

ResMed AirFit F30i Full Face Mask Manual de usuario

Tipo
Manual de usuario
Este manual también es adecuado para