John Bunn DigiO2 Instrucciones de operación

Tipo
Instrucciones de operación
Operation Manual
Important: Do not operate the John Bunn DigiO
2
Finger
Pulse Oximeter without rst reading and understanding
this manual! Save this manual for future use.
JB02007-INS-LAB-RevA07
2 John Bunn DigiO2 Finger Pulse Oximeter
GF, Graham-Field, and John Bunn are registered trademarks of GF Health Products, Inc.
Packaging, warranties, products, and specications are subject to change without notice.
GF Health Products, Inc. is not responsible for typographical errors.
CONTENTS
INTRODUCTION ..................................................................................................................................3
IMPORTANT SAFETY GUIDELINES - PLEASE READ BEFORE USE ...............................................3
INTENDED USE .....................................................................................................................................6
CONTRAINDICATIONS .........................................................................................................................6
DECLARATION OF CONFORMITY: .....................................................................................................6
OVERVIEW ..........................................................................................................................................7
APPEARANCE ......................................................................................................................................7
FEATURES ............................................................................................................................................8
NAME AND MODEL ..............................................................................................................................8
BATTERY INSTALLATION ..................................................................................................................8
OPERATION INSTRUCTIONS ............................................................................................................9
ACCESSORIES (INCLUDED) ...........................................................................................................12
REPAIR AND MAINTENANCE ..........................................................................................................12
MAINTENANCE ...................................................................................................................................12
CLEANING AND DISINFECTION ........................................................................................................13
TROUBLESHOOTING .......................................................................................................................14
SYMBOL KEY ....................................................................................................................................15
TECHNICAL SPECIFICATIONS ........................................................................................................16
LIMITED WARRANTY .......................................................................................................................17
John Bunn DigiO2 Finger Pulse Oximeter 3
INTRODUCTION
Dear User: Please read and understand all instructions before using the John Bunn
DigiO2 Finger Pulse Oximeter, JB02007.
Important Safety Guidelines - Please Read Before Use
The safety statements presented in this chapter refer to the basic safety information
that the operator of the John Bunn DigiO2 Finger Pulse Oximeter shall pay attention
to and abide by. There are additional safety statements in other chapters or sections,
which may be the same as or similar to the following, or specic to the operations.
WARNING: Indicates a potential hazard situation or unsafe practice that, if not
avoided, could result in death or serious injury.
sCaution: Indicates a potential hazard or unsafe practice that, if not avoided, could
result in minor personal injury or product/property damage.
Note: Provides application recommendations or other useful information to ensure that
you get the most from your product.
WARNINGS:
Important! Read and understand this manual before using the oximeter. Do not use this
device without proper instruction from a Healthcare Professional. If the oximeter is not
properly operated, personal injury and/or damage to the oximeter could result.
If components are damaged or missing, contact your dealer immediately. DO NOT use
substitute parts.
4 John Bunn DigiO2 Finger Pulse Oximeter
WARNINGS continued
Inspect the device before use to ensure that there is no visible damage that may affect
patient or users safety or measurement performance. Thereafter, inspect the device at
least once a week.
GF Health Products, Inc. assumes no responsibility for any damage or injury caused by
improper installation or use of this product.
DO NOT open the oximeter housing. It contains no serviceable parts. Service must be
performed by trained, authorized personnel only. Otherwise, device failure and health
hazard may occur.
EXPLOSION HAZARD: Do not use the oximeter in the presence of ammable anesthet-
ics, explosive substances, vapors or liquids.
The oximeter is designed for real-time and rapid measurement. It is not suitable for
long-time continuous patient monitoring. DO NOT apply the sensor to the same nger
for more than two hours. If any readings are found to be abnormal, please change the
position of the sensor.
The oximeter is intended only as an adjunct in patient assessment. It must be used in
conjunction with clinical signs and symptoms. It is intended as a measuring, not a treat-
ment, device.
John Bunn DigiO2 Finger Pulse Oximeter 5
WARNINGS continued
Magnetic and electrical elds are capable of interfering with the proper performance of
the oximeter. For this reason, ensure that all external devices operated in the vicinity of
the oximeter comply with the relevant EMC requirements. Mobile phones, X-ray equip-
ment, and/or MRI devices are a possible source of interference, as they may emit high
levels of electromagnetic radiation.
DO NOT clip this device on edema or tender tissue.
The infrared light emitted from the device (it is invisible) is harmful to the eyes. DO NOT
stare into the light.
Keep this device out of the reach of children.
sCautions
To ensure patient safety, use only parts and accessories specified in this manual.
Keep the oximeter away from dust, vibration, corrosive substances, explosive mate-
rials, high temperature and moisture.
If the oximeter gets wet, stop operation immediately and do not resume operation
until it is dry. When it is carried from a cold environment to a warm and/or humid
environment, allow it to warm to room temperature before using it.
DO NOT use sharp objects to press the front panel button.
DO NOT immerse the oximeter in liquid.
6 John Bunn DigiO2 Finger Pulse Oximeter
sCautions continued
DO NOT subject the oximeter to high temperature or high pressure steam disinfec-
tion. Refer to the REPAIR AND MAINTENANCE/Cleaning and Disinfection section of
this manual for instructions for cleaning and disinfection.
Intended use
The John Bunn DigiO2 Finger Pulse Oximeter, which combines a sensor and monitor
into one device, is used to measure pulse oxygen saturation (SpO2 ) value, pulse rate
value, and blood perfusion index (PI) or pulse strength in home or clinical settings. It is
designed for adult and pediatric patients, and ts ngers of thickness .39" - .87" (10mm
- 22mm).
Contraindications
WARNING: DO NOT use the oximeter while the patient is undergoing MRI or CT scan-
ning.
Declaration of Conformity:
This device complies
with the following
standards:
IEC 60601-1
ISO 9919
ISO 21647
and follows the provisions of the council directive MDD93/42/
EEC
John Bunn DigiO2 Finger Pulse Oximeter 7
OVERVIEW
Appearance
display
key
nameplate
battery
cover
lanyard
hole
display
screen
rubber
cushion
Figure 1
front view Figure 2
rear view
8 John Bunn DigiO2 Finger Pulse Oximeter
Features
The John Bunn DigiO2 Finger Pulse Oximeter provides a simple way to spot-check
patients by combining the sensor and monitor into one integrated, compact, easy to
use device with single-button operation. The oximeter can accurately measure pulse
oxygen saturation (SpO2 ) value, pulse rate value, and blood perfusion index (PI) or
pulse strength. When a nger is inserted into the sensor's rubber cushion, the SpO2
value and pulse rate value automatically display on the screen. The oximeter features
a bright-color LED screen with multi-directional displays that enable the user to read
results from any angle. The oximeter features both audible and visible alarms and a low
voltage indicator. The oximeter powers off automatically in eight seconds when not in
use. The oximeter provides approximately 2,000 spot checks with the use of alkaline
batteries.
Name and Model
Name: John Bunn DigiO2 Finger Pulse Oximeter
Model: JB02007
BATTERY INSTALLATION
1. Refer to Figure 3; insert two AAA batteries, oriented
as shown, into the battery compartment.
2. Replace the cover.
Note: Ensure that the batteries are correctly installed. In-
correct installation will prevent the device from operating.
Figure 3
battery installation
John Bunn DigiO2 Finger Pulse Oximeter 9
OPERATION INSTRUCTIONS
Notes:
Ensure that the finger is dry before being inserted into the oximeter.
Do not shake the finger; keep finger motionless during oximeter use.
Avoid placing the device on a limb which is wrapped with a cuff for blood pressure mea-
surement or undergoing venous infusion.
Ensure that no obstruction (contamination, scar tissue, fingernail polish, acrylic fingernail,
etc.) exists at the site where the sensor is placed; such obstruction could affect the signal
received by the sensor and result in an incorrect measurement.
Vigorous exercise and electrosurgical device interference may affect the measuring ac-
curacy.
1. Open the clip as shown in Figure 4.
2. Insert nger into the clip's rubber cushion (en-
sure that the nger is correctly positioned), and
then gently clip the nger.
3. The device will power on automatically in two
seconds, and start to display the software ver-
sion number.
Figure 4
operation
10 John Bunn DigiO2 Finger Pulse Oximeter
4. The screen will then display the data as
shown in Figure 5. After the waveform
and value become stable, the user can
read the information on the display
screen.
5. When the screen displays as shown in
Figure 5, press the Display Key to change
the display screen's orientation:
a. Press the Display Key once; the
display screen will rotate 180° and
display as shown in Figure 6.
b. Press the Display Key twice; the
screen will now display as shown in
Figure 7.
c. Press the Display Key three times;
the display screen will rotate 180°
and display as shown in Figure 8.
d. Press the Display Key four times; the display screen will return to the original
screen as shown in Figure 5.
Figure 5
Figure 6
Figure 7
Figure 8
John Bunn DigiO2 Finger Pulse Oximeter 11
6. Press the Display Key for approximately
two seconds; the screen will display as
shown in Figure 9. Differences between
Figure 9 and Figure 5 are as follows:
a. In Figure 5, SpO2 and PR (pulse rate)
are being monitored and displayed
on the screen.
b. In Figure 9, SpO2 and PI (blood perfu-
sion index) are being monitored and
displayed on the screen.
7. Press the Display Key twice; the screen will now display as shown in Figure 10.
8. The Display screen as shown in Figure 9 or Figure 10 will return to the original
screen as shown in Figure 5 or Figure 6 if without operation in ten seconds.
9. Alarm Indicator: When measuring, if SpO2 value and pulse rate value are outside the
preset alarm limits, the device will alarm automatically and the value which is out-
side the limit will ash on the screen; press the Display Key to suspend the alarm
beep (see alarm table below).
Alarm SpO2Outside limit alarm: three
beeps
Preset limit: ≤90%
Pulse
Rate
Outside limit alarm: two beeps Preset limits: Low ≤50bpm, High ≥120bpm
Figure 9
Figure 10
12 John Bunn DigiO2 Finger Pulse Oximeter
ACCESSORIES (INCLUDED)
1 One lanyard
2 Two AAA alkaline batteries
3 One pouch
4 One operation manual
5 One quality certicate
REPAIR AND MAINTENANCE
WARNING: DO NOT open the oximeter housing. It contains no serviceable parts.
Service must be performed by trained, authorized personnel only. Otherwise, device
failure and health hazard may occur.
Maintenance
In order to ensure the oximeter's longevity and continued dependable operation,
please observe the following maintenance recommendations.
Change the batteries when the low-voltage indicator displays.
Clean the surface of the device before using: Wipe the device with alcohol, then let it air dry or wipe it
dry. Repeat as necessary.
Take out the batteries if the oximeter will not be used for a long period of time.
Keep the device in a dry environment. Humidity may shorten the life of the device, or even damage it.
John Bunn DigiO2 Finger Pulse Oximeter 13
Environmental requirements Operation Temperature: 41°F ~104°F (5°C ~40°C)
Relative humidity: 30%~80%
Atmospheric pressure: 10.15 psi~15.37 psi (70kPa~106kPa)
Storage Temperature: -4°F to 131°F (-20ºC to 55ºC)
Relative humidity: < 95%
Cleaning and Disinfection
1. Do not immerse the device in liquid. Clean the device only by using a surface-clean
technique: Wipe with a clean, soft cloth saturated with a solution of 75% isopropyl
alcohol.
If low-level disinfection is required, wipe with a 1:10 bleach to water solution.
2. Wipe with a clean, soft cloth saturated with clean water and air dry, or dry with a
clean, soft cloth.
sCaution: Do not sterilize with irradiation, steam, or ethylene oxide.
sCaution: Do not disinfect with high temperature or high pressure steam.
sCaution: Do not use the sensor if it is damaged.
14 John Bunn DigiO2 Finger Pulse Oximeter
TROUBLESHOOTING
Problem Possible Reason Solution
The SpO2 and Pulse Rate
display is unstable
1. The nger is not correctly
positioned in the device
1. Position the nger correctly
and try again
2. The nger is shaking or the
patient is moving
2. Ensure that patient is able to
remain immobile
The device does not turn
on
1. The batteries are low 1. Change the batteries
2. The batteries are not correctly
installed
2. Reinstall batteries
3. The device has malfunctioned 3. Contact your Graham-Field
distributor
The indicator light
suddenly turns off
1. The device powers off
automatically when it gets no
signal for eight seconds
1. Normal
2. The batteries are almost
drained
2. Change the batteries
John Bunn DigiO2 Finger Pulse Oximeter 15
SYMBOL KEY
Symbol Description
Type BF
Warning: See operation manual
The pulse oxygen saturation
PI Perfusion Index or Pulse Strength
Pulse rate (beats per minute)
Low battery voltage
SN Serial number
Do not treat this device as household waste—
separate collection required
16 John Bunn DigiO2 Finger Pulse Oximeter
TECHNICAL SPECIFICATIONS
Display mode LED Display
Power supply requirement Two 1.5V AAA alkaline batteries
Supply voltage: 2.7V~3.3V
Operating current ≤40mA
SpO2 Parameter
Specications
Measuring range: 35%~99%
Accuracy: ±2% during 75%~99%, ±3% during 50%~75%
Pulse Rate Parameter
Specications
Measuring range: 25-250 bpm
Accuracy: ±2 bpm or ±2%
Pulse Strength Parameter
Specications:
Measuring range: 0.2-20%
Accuracy: ±0.1% during 0.2%~2%, ±1% during 2%~10%,
±2% during 10%~20%
Resistance to interference of
surrounding light
The difference between the value measured in the condition of indoor
natural light and that of darkroom is less than ±1%
Resistance to interference of
man-made light
Values of SpO2 and Pulse Rate can be accurately measured by pulse
oxygen simulator
Dimensions (L x W x H) 2.59" x 1.42" x 1.3" (66 mm x 36 mm x 33 mm)
John Bunn DigiO2 Finger Pulse Oximeter 17
TECHNICAL SPECIFICATIONS continued
Net weight with batteries 2 oz. (60g)
Classication The type of protection against electric shock: Internally powered
equipment
The degree of protection against electric shock: Type BF applied part
The degree of protection against harmful ingress of liquids: Ordinary
equipment without protection against ingress of water
Electro-Magnetic Compatibility: Group I, Class B
LIMITED WARRANTY
GF Health Products, Inc. warrants the John Bunn DigiO2 Finger Pulse Oximeter, Model
JB02007, for a period of one year for defects in workmanship and materials. During the
warranty period, defective items will be repaired or replaced at manufacturer's option
at no charge.
USA, Corporate Headquarters:
Graham-Field Health Products
2935 Northeast Parkway
Atlanta, Georgia 30360
telephone: 800-347-5678, 770-447-1609
fax: 800-726-0601, 678-291-3232
www.grahameld.com
© 2007 GF Health Products, Inc.
JB02007-INS-LAB-RevA07
JB02007-INS-LAB-RevA07
Manual de Operación
Importante: No debe operar el John Bunn DigiO
2
Oxímetro sin leer y entender este manual!
Guarde este manual para uso en el futuro.
2 John Bunn DigiO2 Oxímetro de Pulso
GF, Graham-Field y John Bunn son marcas registradas de GF Health Products, Inc.
El embalaje, garantías, productos y specicaciones son sujetos a cambiar sin aviso.
GF Health Products, Inc. no es responsable por errores tipográcos.
CONTENIDOS
INTRODUCCIÓN ............................................................................................................................................3
DIRECTIVAS IMPORTANTES DE SEGURIDAD – POR FAVOR DE LEER ANTES DE USO .....................3
USO ENTENDIDO .................................................................................................................................6
CONTRAINDICACIONES ......................................................................................................................6
DECLARACIÓN DE CONFORMIDAD: .................................................................................................6
GENERALIDADES ........................................................................................................................................7
APARIENCIA .........................................................................................................................................7
CARACTERÍSTICAS .............................................................................................................................8
NOMBRE Y MODELO ...........................................................................................................................8
INSTALACIÓN DE LAS BATERÍAS ..............................................................................................................8
INSTRUCCIONES DE OPERACIÓN .............................................................................................................9
ACCESORIOS (INCLUIDOS) ......................................................................................................................12
REPARACIÓN Y MANTENIMIENTO ...........................................................................................................12
MANTENIMIENTO ...............................................................................................................................12
LIMPIAR Y DESINFECTAR .................................................................................................................13
DETECTANDO PROBLEMAS .....................................................................................................................14
SÍMBOLOS ..................................................................................................................................................15
ESPECIFICACIONES TÉCNICAS ...............................................................................................................16
GARANTÍA ...................................................................................................................................................17
John Bunn DigiO2 Oxímetro de Pulso 3
INTRODUCCIÓN
Estimado usuario: Por favor de leer y entender las instrucciones antes de usar el John
Bunn DigiO2 Oxímetro de Pulso de uso Dactilar JB02007. Guarde estas instrucciones
para uso en el futuro.
DIRECTIVAS IMPORTANTES DE SEGURIDAD – POR FAVOR DE LEER
ANTES DE USO
Las declaraciones de seguridad presentado en este capitulo reeren a la información
de seguridad básica que el operador del John Bunn DigiO2 Oxímetro de Pulso de uso
Dactilar debe ponerle atención y seguirlo. Hay declaraciones de seguridad adicionales
en otros capítulos o secciones, cuales pueden ser lo mismo o similar a lo siguiente o
especico a la operación.
Advertencia: Indica una situación de peligro o una practica inseguro que, so no es
evitado, puede resultar en muerte o heridas serias.
sCaución: Indica una situación de peligro o una practica inseguro que, so no es
evitado, puede resultar en herida leve o daño al producto/propiedad.
Aviso: Proporcione recomendaciones para la aplicación y otra información útil para asegu-
rar que obtengas lo más que puedes del producto.
Advertencias:
Importante! Lee y entiende este manual antes de usar el oxímetro. No debe usar este
aparato sin instrucciones propias de un Profesional de Cuido de Salud. Si el oxímetro no
es operado correctamente, herida personal y/o daño al oxímetro puede resultar.
4 John Bunn DigiO2 Oxímetro de Pulso
Advertencias continuadas:
Si los componentes son dañados o si le falta, llame a su distribuidor inmediatamente.
NO DEBE usar partes sustitutos.
Es necesario inspeccionar el aparato antes de uso para asegurar que no hay daño visi-
ble que puede afectar al paciente o la seguridad del usuario o actuación de medida. A
partir de entonces, debe inspeccionar el aparato por lo menos una vez a la semana.
GF Health Products, Inc. no asume ninguna responsabilidad de daño o lesión causado
por instalación, asamblea o uso impropio de este aparato.
No debe usar este aparato si está dañado o se le falta componentes.
NO DEBE abrir la carcasa del oxímetro. No contiene partes servible. Servicio debe ser
hecho por solamente por un ingeniero de servicios cualicado. Los usuarios no son
permitidos a mantenerlos ellos propios. Fallo del aparato y peligro puede ocurrir.
PELIGRO DE EXPLOSIÓN: No debe usar el oxímetro en la presencia de un ambiente con
gas inamable como agentes anestesia, substancias explosivas, vapores o líquidos que
se pueden encender.
El oxímetro es designado para medidas rápidas. No es apto para monitor en continuo
por largo tiempo al paciente. NO DEBE aplicar el censor al mismo dedo por más de dos
horas. Si los resultas son anormales, por favor de cambiar la posición del censor. El
oxímetro es entendido solamente como un adjunto en la evaluación del paciente. Debe
ser usado en conjunción con señales clínicas y síntomas. Es entendido para un aparato
de medir no tratamiento.
John Bunn DigiO2 Oxímetro de Pulso 5
Advertencias continuadas:
Campos eléctricas y magnético son capaces de interferir con la prestación correcto del
oxímetro. Por esta razón, asegure que todos los aparatos externos operados en la vecin-
dad del oxímetro cumplen con los requisitos de EMC. Celulares, equipo de radiografía, y
aparatos de MRI pueden interferir porque emiten niveles altos de radiación electromag-
néticos.
NO DEBE ser puesto en un edema o tejido tierno.
La luz (la luz infrarrojo es invisible) emitido por el aparato le puede hacer daño a los ojos
así que no debe jarse en la luz.
NO DEBE jarse directamente a la luz.
Este aparato debe estar guardado fuera del alcance de niños.
sPrecauciones:
Para asegurar la seguridad del paciente, use partes y accesorios solamente especifi-
cados en este manual.
Guarde al oxímetro lejos de polvo, vibración, sustancias corrosivas, materiales explo-
sivas, temperaturas altas y humedad.
Si el oxímetro se moje, por favor de parar de usarlo. Cuando se mueve de un am-
biente frió a un ambiente caliente y con humedad, por favor de no usarlo inmedia-
tamente. Cuando el aparato se cambia de un ambiente frió a uno caliente o húmedo,
permite que se caliente a la temperatura del cuarto antes de usarlo.
NO DEBE apretar el botón al frente con objetos afilados.
6 John Bunn DigiO2 Oxímetro de Pulso
sPrecauciones continuadas:
NO DEBE sumergir el oxímetro en líquido.
Temperatura alta o desinfección usando vapor de alta presión no es permitido para
el oxímetro. Refiere al manual para instrucciones para limpiar y desinfección. Refie-
re a la sección de Reparación y Mantenimiento / Limpiando y Desinfectando de este
manual para instrucciones para limpiar y desinfectar.
Uso Entendido
El John Bunn DigiO2 Oxímetro de Pulso de uso Dactilar, que combine un censor y moni-
tor en un aparato entendido para medir el valor de la saturación del pulso de oxígeno
(SpO2), el valor de frecuencia de pulso y índice de perfusión de sangre (PI) o fuerza de
pulso en escenarios de casa y clínicas . Es deseando para adultos y pacientes pediátri-
cos, y cave en dedos que son grueso de .39" - .87" (10mm – 22mm).
Contraindicaciones
Advertencia: NO DEBE usar el oxímetro cuando el paciente está debajo un MRI o
exploración de CT.
Declaración de Conformidad:
El fabricante declare que este aparato
cumple con los siguientes estándares:
IEC 60601-1
ISO 9919
ISO 21647
Y sigue las provisiones del concejo directiva MDD93/42/EEC
John Bunn DigiO2 Oxímetro de Pulso 7
GENERALIDADES
Apariencia
botón
de mostra
placa
de nombre
protector
de batería
agujero
del cordón
pantalla
de mostra
cojín de goma
Figura 1
vista frontal Figura 2
vista de atrás
8 John Bunn DigiO2 Oxímetro de Pulso
Características
Este John Bunn DigiO2 Ometro de Pulso de uso Dactilar provee un todo simple para
chequear a los pacientes combinando un censor y monitor en un aparato integrado y com-
pacto que es fácil de usar y que tiene operacn de un botón. El John Bunn DigiO2 Ometro
de Pulso de uso Dactilar puede ser usado para medir el valor de la saturación del pulso
de oxígeno (SpO2), el valor de frecuencia de pulso y índice de perfusn de sangre (PI) o
fuerza de pulso. En poniendo su dedo en el censor, el valor de SpO2 y de frecuencia de pulso
se vera en la pantalla autoticamente. El oxímetro tiene una pantalla de LED con color
brillante que da más colores vibrantes y tiene una vida de operacn mas largo. La pantalla
es de dirección ltiple para que el usuario pueda leer los resultados de cual quiere ángulo.
El oxímetro tiene alarmes audibles y visuales y tambn tiene un indicador de voltaje baja.
El oxímetro se apagara automáticamente en 8 segundos cuando no es en uso. El oxímetro
provee aproximadamente 2000 inspecciones en el alto usando baterías de alcalina.
Nombre y Modelo
Nombre: John Bunn DigiO2 Oxímetro de Pulso
Modelo: JB02007
INSTALACIÓN DE LAS BATERÍAS
1. Reere a la Figura 3; Inserta dos AAA baterías en el com-
partimiento correctamente en su propia dirección.
2. Coloque el protector.
Aviso: Asegurase instalar las baterías correctamente. Instala-
ción incorrecta puede hacer que el aparato no trabaja.
Figura 3
instalación de las baterías
John Bunn DigiO2 Oxímetro de Pulso 9
INSTRUCCIONES DE OPERACIÓN
Avisos:
Asegure que el dedo está seco antes de insertarlo en el oxímetro.
El dedo debe estar seco antes de ponerlo correctamente en el oxímetro.
No debe hacer temblar su dedo cuando estás usando el aparato.
Evite poner el aparato en el mismo brazo donde tiene el medidor de presión o cuando se
está siendo un infusión venosa.
No permite que ningún obstrucción (contaminación, cicatriz, esmalte, unas acrílicas etc.)
esté en el sitio donde está colocado el censor. Estos obstrucciones pueden afectar la señal
recibidos por el censor y puede resultar en una medida incorrecta.
Ejercicio vigoroso y interferencia de un aparato electro quirúrgica pueden afectar la preci-
sión de medir.
1. Abre el aparato como se ve en la Figura 4.
2. Mete su dedo en el cojín de goma (asegurase que
su dedo está en la posición correcta).
3. El aparato se aprenderá automáticamente en 2
segundos, y le enseñará el número de la versión de
software.
Figura 4
operación
10 John Bunn DigiO2 Oxímetro de Pulso
La pantalla le enseñará los datos (como en la
Figura 5). El usuario puede obtener la informa-
ción directamente de la pantalla después de
que se establece el valor.
4. La pantalla le enseñará los datos (como
en la Figura 5). El usuario puede obtener
la información directamente de la pantalla
después de que se establece el valor.
5. Cuando la pantalla se ve como en la Figura
5, apreté el Botón de Mostrar para cambiar
la pantalla:
a. Apreté el Botón de Mostrar una vez, la
pantalla se volteará 180° grados (como
se ve en la Figura 6).
b. Apreté el Botón de Mostrar dos veces,
la pantalla se cambiara a la Figura 7.
c. Apreté el Botón de Mostrar tres veces, la pantalla (que se ve como en la Figura 8)
se volteará 180° grados.
d. Apreté el Botón de Mostrar cuatro veces, la pantalla se volviera a la pantalla
original como en la Figura 5.
Figura 5
Figura 7
Figura 8
John Bunn DigiO2 Oxímetro de Pulso 11
6. Apriete el Bon de Mostrar por un periodo
mas largo (2 segundos), la pantalla se ve
como en la Figure 9. Las diferencias entre
Figura 9 y la Figura 5 son los siguientes:
a. En la Figura 5, el valor de la saturación
del pulso de oxígeno (SpO2) y el valor
de frecuencia de pulso son observados
en la pantalla.
b. En la Figura 9, el valor de la saturación del pulso de oxígeno (SpO2) y índice de
perfusión de sangre (PI) o fuerza de pulso son observados en la pantalla.
7. Apreté el Botón de Mostrar dos veces, la pantalla se cambiara a la Figura 10.
8. La pantalla que se ve en la Figura 9 o Figura 10 se volverá a la pantalla original como
se ve en la Figura 5 o Figura 6 si no hay operación en 10 segundos.
9. Indicador de Alarma: Cuando estás mediando, si el valor de la saturación del pulso de
ogeno y frecuencia de pulso sobrepasa los limites del alarme, el aparato automática-
mente pone el alarme y el valor en que excede se vera en la pantalla; en este tiempo
apreté el Botón de Mostrar para suspender el sueno del alarme (vea la tabla de alarmes).
Limites de
los alarmes
presentes
Alarme de SpO2Limite inferior (3
pitidos)
Alarme presente: <90%
Alarme de frecuen-
cia de pulso
Limite alta (2 pitidos) Alarmes presente: Bajo <50bpm, Alto >120bpm
Figura 9
12 John Bunn DigiO2 Oxímetro de Pulso
ACCESORIOS (INCLUIDOS)
1 Cordón
2 AAA Baterías de alcalina
3 Una Bolsita
4 Un Manual de Operación
5 Un Certicado de Calidad
REPARACIÓN Y MANTENIMIENTO
Advertencia: NO DEBE abrir la carcasa del oxímetro. No contiene partes servibles.
Servicio debe ser hecho solamente por un personal autorizado. Si no, falla del aparato
y daño a la salud puede ocurrir.
Mantenimiento
Para asegurar la longevidad y operación continuado seguro del oxímetro, por favor
observe las siguientes recomendaciones para mantenimiento.
Cambie las baterías cuando el indicador del voltaje bajo que se ve en la pantalla.
Limpie la supercie del aparato antes de uso. Limpie el aparato con alcohol primero, déjalo que se seca y
límpialo hasta que está seco. Repito si es necesario.
Saque las baterías del oxímetro si no va usar el aparato por un periodo de tiempo largo.
Guarde el aparato en un ambiente seco. Humedad puede reducir o dañar la vida del aparato.
John Bunn DigiO2 Oxímetro de Pulso 13
Ambientes requeridos Operación Temperatura: 41°F ~104°F (5°C ~40°C)
Humedad relativo: 30%~80%
Presión atmosférico: 10.15psi~15.37psi (70kPa~106kPa)
Almacenaje Temperatura: -4°F - 131°F (-20ºC- 55ºC))
Humedad relativa: < 95%
Limpiar y Desinfectar
1. No debe sumergir este aparato en líquidos. Limpia solamente con la técnica de lim-
piar supercie. Debe usar la gasa suave y saturarlo con la solución de 75% isopropí-
lico alcohol. Si se necesita desinfectar de un nivel baja, usa una solución de lejía de
1:10.
2. Limpie con un paño suave y limpio saturado en agua limpia. Sécalo con un paño
seco y limpio.
sCaución: No debe Esterilizar por vapor irradiación óxido de etileno.
sCaución: No debe desinfectar con temperatura alto o vapor de alta presión.
sCaución: No debe usar el censor si está dañado.
14 John Bunn DigiO2 Oxímetro de Pulso
DETECTANDO PROBLEMAS
Problema Razón Posible Solución
El SpO2 y Frecuencia
de Pulso no son
inestable
1. El dedo no está puesto correctamente
en el aparato
1. Coloque el dedo correctamente y
trato de nuevo
2. El dedo está temblando o el paciente
está moviendo
2. Asegure que el paciente se quede
inmóvil
El aparato no se
prende
1. Las baterías están bajas 1. Cambie las baterías
2. Las baterías no están puestas correc-
tamente
2. Pon las baterías de nuevo
3. El aparato se fallo 3. Llame a su distribuidor de
Graham-Field
La luz del indicador se
apaga de repente
1. El aparato se apaga automáticamente
cuando no hay señal por ocho segundos
1. Normal
2. Las baterías se han gastado 2. Cambie las baterías
John Bunn DigiO2 Oxímetro de Pulso 15
SÍMBOLOS
Símbolo Descripción
Tipo BF
Advertencia: Vea el Manual
La saturación de oxígeno de pulso
PI Índice de Perfusión/Intensidad de Pulso
Frecuencia de Pulso (latos por minuto)
Voltaje baja de las baterías
SN Numero Serial
No debe tratar este aparato como residuos de
domésticos—colección separado es necesario
16 John Bunn DigiO2 Oxímetro de Pulso
ESPECIFICACIONES TÉCNICAS
Modo de Mostrar Pantalla de LED
Requisito de suministro eléctrico Dos 1.5V AAA baterías de alcalino
Voltaje de suministro: 2.7V~3.3V
Corriente Operando ≤40mA
Especicaciones Parámetro de
SpO2
Rango de Medida: 35%~99%
Presión: ±2% (durante 75% - 99%), ±3% (durante 50% - 75%)
Especicaciones de Frecuencia de
Pulso
Rango de Medida: 25-250bpm
Presión: ±2bpm o ±2%
Especicaciones de Parámetro de
Intensidad de Pulso
Rango de Medida: 0.2-20%
Presión: ±0.1% (durante 0.2 - 2%), ±1% (durante 2 - 10%), ±2% (du-
rante 10%- 20%)
Resistencia a interferencia de la luz
circundante
La diferencia entre el valor medido en la condición de la luz de inte-
rior natural y de un cuarto oscuro es menos de ±1%.
Resistencia a interferencia de luz
articial
Valores de SpO2 y Frecuencia de Pulso pueden ser medidos con
presión por un simulador de oxígeno de pulso
Dimensiones (L x W x H) 2.59" x 1.42" x 1.3" (66mm x 36mm x 33mm)
John Bunn DigiO2 Oxímetro de Pulso 17
Especicaciones Técnicas Continuadas
Peso Neto 2 onzas. (60g) con Baterías
Clasicación El tipo de protección contra descarga eléctrica: Equipo con motor
interior
El grado de protección contra descarga eléctrica: Parte aplicado de
Tipo BF
El grado de protección contra un inmersión de líquidos dañinos:
Equipo ordinario sin protección contra inmersión de agua
Compatibilidad Electromagnético: Grupo I, Clase B
GARANTÍA
GF Health Products, Inc. garantía el John Bunn DigiO2 Oxímetro de Pulso, Modelo
JB02007 por un periodo de un año por defectos en la factura y materiales. Durante el
periodo de garantía, artículos defectivos pueden ser arreglados y cambiadas por el
fabricante a no precio.
USA, Jefaturas Corporativas :
Graham-Field Health Products
2935 Northeast Parkway
Atlanta, Georgia 30360
tel: 800-347-5678, 770-447-1609
fax: 800-726-0601, 678-291-3232
www.grahameld.com
© 2007 GF Health Products, Inc.
JB02007-INS-LAB-RevA07
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36

John Bunn DigiO2 Instrucciones de operación

Tipo
Instrucciones de operación

en otros idiomas