Intended Use
The Tango CPAP system is indicated for the treatment of obstructive sleep apnea (OSA) in
patients weighing more than 66 lb (30 kg). It is intended for home and hospital use.
CAUTION
Federal law restricts this device to sale by or on the order of a physician.
Contraindication
Positive airway pressure therapy may be contraindicated in some patients with the
following pre-existing conditions:
severe bullous lung disease
pneumothorax
pathologically low blood pressure
dehydration
cerebrospinal fluid leak, recent cranial surgery or trauma.
Adverse Effects
Patients should report unusual chest pain, severe headache or increased breathlessness
to their prescribing physician. An acute upper respiratory tract infection may require
temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the Tango:
drying of the nose, mouth or throat
nosebleed
bloating
ear or sinus discomfort
eye irritation
skin rashes.
Parts of the Tango
Your Tango comprises:
C-Series Tango flow generator (shown below)
Power cord
6’6” (2 m) air tubing.
Signal cable socket
(humidifier use only)
Air outlet
Power cord socket
Air inlet
(contains air filter)
2
Masks
The ResMed mask systems recommended for use with the Tango are:
Mirage™ Full Face Mask Series II Meridian™ Nasal Mask
Ultra Mirage™ Full Face Mask Mirage™ Nasal Mask
Mirage Swift™ Nasal Pillows System Mirage Activa™ Nasal Mask
Mirage Swift II Nasal Pillows System Mirage Vista™ Nasal Mask
Papillon™ Nasal Mask Ultra Mirage™ II Nasal Mask
Silent Papillon™ Nasal Mask Mirage Liberty™ Full Face Mask
Mirage Quattro™ Full Face Mask
For information on using masks, see your mask manual.
Humidifiers
For information on using the ResMed C-Series Heated Humidifier or the ResMed
Passover, see your humidifier manual.
The Tango Interface
Heating indicator
Power indicator
Standby mode When the Tango is powered, the power indicator is lit and the run hours are displayed on
the LCD. All changes to altitude and ramp settings are made in standby mode.
Ramp Setting a ramp time enables therapy to start at a low pressure and build up to full
pressure during the ramp time (see “Set ramp time”).
Altitude Changes in altitude affect the pressure delivered by the Tango. The altitude feature
compensates for pressure changes related to changes in altitude (see “Set altitude”).
Run hours Run hours are displayed when the Tango is in standby mode.
Humidifier The humidifier buttons are used when using the C-Series Heated Humidifier and when
setting the altitude. The heating indicator lights when the humidifier is on.
3
English
Setting up the Tango
CAUTION
Be careful to place the device where it cannot be bumped, and where no one will
trip over the power cord.
If you put the device on the floor, make sure the area is free from dust and clear
of bedding, clothes or other objects that could block the air inlet.
Make sure the area around the device is dry and clean.
WARNING
Make sure the power cord and plug are in good condition and the equipment is
not damaged.
Only ResMed air tubing should be used with the device. A different type of air
tubing may alter the pressure you actually receive and reduce the effectiveness
of your treatment.
Blocking the hose and/or air inlet of the device while in operation could lead to
overheating of the device.
1
3
4
2
Connect the power cord to the rear of the Tango.
Plug the other end into a power outlet.
Connect the air tubing to the air outlet of the Tango.
Connect the assembled mask system to the free end of the air tubing.
1.
2.
3.
4.
4
Using the Tango
Start therapy
Press
when the Tango is in standby mode. A circulating dash symbol displays on the
LCD.
If ramp is set, the ramp time displays on the LCD. When ramping is finished, therapy
continues at the prescribed pressure.
If no ramp is set, therapy starts immediately at the prescribed pressure.
Stop therapy
Press
.
Set ramp time
Repeatedly press
until the LCD displays the desired time.
Press
to accept changes and start therapy. (The Tango returns to standby mode
after 5 seconds if no button is pressed.)
The ramp settings are (in minutes): Off, 10, 20, 45.
1.
2.
Set altit
ude
Hold
down for 3 seconds until the current altitude setting is displayed.
Press
or to change the altitude setting.
Press
to accept changes and return to standby mode. (The Tango returns to
standby mode after 10 seconds if no button is pressed.)
The altitude settings are:
1 (0–2000’/0–609 m),
2 (2001–40
00’/610–1219 m),
3 (4001–6000’/1220–1828 m),
4 (6001–8500’/1829–2591 m).
1.
2.
3.
Cleaning and Maintenance
WARNING
Beware of electric shock. Do not immerse the flow generator or power cord in water.
Always unplug the flow generator before cleaning and make sure it is dry before
reconnecting.
CAUTION
Do not hang the air tubing in direct sunlight as it may harden over time and
eventually crack.
Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing
or antibacterial soaps or scented oils to clean the air tubing or the Tango. These
solutions may cause hardening and reduce the life of the product.
Daily: Disconnect the air tubing and hang it in a clean, dry place until next use.
Weekly:
Remove the air tubing and wash it in warm water and mild detergent. Rinse
thoroughly, hang and allow to dry.
Before next use, reassemble the mask and headgear.
Reconnect the air tubing.
Monthly:
Clean the exterior of the Tango with a damp cloth.
Check the air filter for holes and blockage by dirt. Replace it every six months, or more
often in a dusty environment.
For information on cleaning your mask or humidifier, refer to the manual supplied with
your mask or humidifier.
1.
2.
3.
5
English
Changing the Air Filter
Air inlet
Air filter
CAUTION
Do not wash the air filter.
With the power cord and air tubing disconnected, turn the Tango upside down so the
air inlet is facing you. The air filter is inside the air inlet.
Pull the old air filter out of the air inlet.
Put the new air filter into the air inlet in the direction shown by the arrow, with the
blue side facing towards you.
Troubleshooting
Problem Possible Cause Solution
Tango won’t start Power cord not connected
properly.
Check power cord.
Faulty power outlet. Try another power outlet.
No/low air flow The air tubing is not connected
properly.
Adjust the air tubing connection.
The air tubing is blocked,
pinched or punctured.
Unblock or free the air tubing. Check the
air tubing for punctures.
Treatment pressure
seems either low or too
high
Altitude setting incorrect. Adjust altitude setting.
LCD displays an error
message and treatment
stops or can’t be started
Problem with motor. Contact your supplier.
1.
2.
3.
6
Technical Specifications
Operating pressure range 4 to 20 cm H
2
O
Dimensions (L x W x D) 8.5” x 7.4” x 4.5”
(215 mm x 189 mm x
114 mm)
Weight 2.4 lb (1.1 kg)
Power supply Input range
110V, 60Hz, 20VA/10W (typical pow
er consumption),
36VA/15W (maximum power consumption)
Operating temperature +41°F to +104°F (+5°C to +40°C)
Operating humidity 10–95% non-condensing
Storage and transpor
t
temperature
-
4°F to +140°F (-20°C to +60°C)
Storage and transport
humidity
10–95% non-condensing
Atmospheric pressure range 1060 hP
a (sea lev
el) to 680 hPa
(8500’/ 2591 m)
Electromagnetic
Compatibility
P
roduct complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial and light industry environments
Air Filter Two-lay
ered, powder-bonded, polyester non-woven fiber
Air Tubing Flexible plastic, 6’6” (2 m)
Air Outlet T
he 0.87” (22 mm) conical air outlet complies with EN 1281-1
IEC 60601-1 Classification Class II (double insulation), T
ype CF
The manufacturer reserves the right to change these specifications without notice.
Symbols which appear on the Tango
Attention. Consult accompanying documents; Type CF equipment; Class II
equipment;
IPX1
Drip proof; Canadian Standards Association.
7
English
General Warnings and Cautions
A warning alerts you to possible injury.
Read the entire guide before using the Tango.
The advice in this guide should not supersede instructions given by the prescribing
physician.
The Tango should only be used with masks (and connectors*) recommended by
ResMed, or by a physician or respiratory therapist. A mask should not be used unless
the Tango is turned on and operating properly. The vent hole or holes associated with
the mask should never be blocked.
Explanation: The Tango is intended for use with special masks (or connectors
*
) with
vent holes allowing a continuous flow of air out of the mask.
When the device is turned on and functioning properly, new air from the device
flushes the exhaled air out through the mask vent holes. If the device is not operating,
insufficient fresh air is provided through the mask, and the exhaled air may be
rebreathed. Rebreathing exhaled air for longer than several minutes can sometimes
lead to suffocation. If there is a power failure or the device machine malfunctions,
remove the mask.
The Tango can be set to deliver pressures up to 20 cm H
2
O. In the unlikely event of
certain fault conditions, pressures up to 30 cm H
2
O are possible.
If oxygen is used with this device, the oxygen flow must be turned off when the
device is not operating. If oxygen has been left on while the device is not operating,
disconnect the oxygen and wait 30 minutes before turning on the flow generator
again.
Explanation: When the CPAP device is not in operation and the oxygen flow is left on,
oxygen delivered into the air delivery tubing may accumulate within the CPAP machine
enclosure and create a risk of fire.
Oxygen supports combustion. Do not use oxygen while you are smoking or in the
presence of an open flame.
Always ensure air is flowing from the device before turning on the oxygen supply.
Always turn the oxygen supply off before stopping the airflow from the device.
Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration
will vary, depending on where the oxygen is introduced, pressure settings, patient
breathing pattern, mask selection and leak rate.
Explosion hazard—do not use in the vicinity of flammable anesthetics.
Do not use the Tango if there are obvious external defects or unexplained changes in
performance.
Do not open the Tango case. There are no user serviceable parts inside. Repairs and
internal servicing should only be performed by an authorized service agent.
A caution explains special measures for safe and effective use of the device.
At low pressures, the flow through the exhalation ports of your mask may not clear all
exhaled gas from the tubing. Some rebreathing may occur.
The airflow for breathing produced by this device can be as much as
11°F (6°C) higher
than the temperature of the room. Be careful if the room temperature is warmer than
90°F (32°C).
* Ports may be incorporated into the mask or in connectors that are near the mask.
8
Servicing
The Tango should be inspected by an authorized ResMed service center 5 years from
the date of manufacture. Before this, the device is intended to provide safe and reliable
operation if it is operated and maintained according to the instructions provided by
ResMed. Warranty details are provided with the device at the time of original supply.
As with all electrical devices, if any irregularity becomes apparent, you should have the
device inspected by an authorized ResMed service center.
Limited Warranty
ResMed warrants that your ResMed product shall be free from defects in material and
workmanship for the period specified below from the date of purchase by the initial
consumer. This warranty is not transferable.
Product Warranty Period
ResMed humidifiers, ResControl™, ResLink™,
ResTraxx™
1 Year
ResMed flow generators 2 Y
ears
Accessories, water chambers and mask systems
(including mask frame, cushion, headgear and tubing).
Excludes single-use devices.
90 Days
Note: Some models are not available in all regions.
If the product fails under conditions of normal use, ResMed will repair or replace, at its
option, the defective product or any of its components. This Limited Warranty does not
cover:
a) any damage caused as a result of improper use, abuse, modification or alteration of the
product;
b) repairs carried out by any service organization that has not been expressly authorized
by ResMed to perform such repairs;
c) any damage or contamination due to cigarette, pipe, cigar or other smoke;
d) any damage caused by water being spilled on or into a flow generator.
Warranty is void on product sold, or resold, outside the region of original purchase.
Warranty claims on defective product must be made by the initial consumer at the point
of purchase.
This warranty is in lieu of all other express or implied warranties, including any implied
warranty of merchantability or fitness for a particular purpose. Some regions or states do
not allow limitations on how long an implied warranty lasts, so the above limitation may
not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed to
have occurred as a result of the sale, installation or use of any ResMed product. Some
regions or states do not allow the exclusion or limitation of incidental or consequential
damages, so the above limitation may not apply to you. This warranty gives you specific
legal rights, and you may also have other rights which vary from region to region.
For further information on your warranty rights, contact your local ResMed dealer or
ResMed office.
9
English
Guidance and Manufacturers Declaration - Electromagnetic Emissions and
Immunity
Guidance and manufacturer’s declaration—electromagnetic emissions
The C-Series Tango system is intended for use in the electromagnetic environment specified below. The
customer or the user of the C-Series Tango system should assure that the device is used in such an environment.
Emissions test Compliance Electromagnetic environment—guidance
RF emissions CISPR11 Group 1 The C-Series Tango system uses RF energy only f
or its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The C-Series Tango system is suitable f
or use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage network that
supplies buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/Flic
ker
Emissions IEC 61000-3-3
Complies
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to EMC information provided in this document.
W
arnings: The C-Series Tango system should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the C-Series Tango system should be observed to verify normal operation
in the configuration in which it will be used.
The use of accessories (eg, humidifiers) other than those specified in this manual is not recommended. They may
result in increased emissions or decreased immunity of the C-Series Tango system.
10
Guidance and manufacturer’s declaration—electromagnetic immunity
The C-Series Tango system is intended for use in the electromagnetic environment specified below. The
customer or the user of the C-Series Tango system should assure that the device is used in such an environment.
Immunity test IEC60601-1-2 test level Compliance level Electromagnetic
environment—
guidance
Electrostatic discharge
(ESD)
IEC 6
1000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be w
ood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for po
wer supply
lines
±2 kV Mains power quality
should be that of a
typical commercial or
hospital environment.
±1 kV for input/output
lines
Not Applicable
Surge
IEC 61000-4-5
±1 kV dif
ferential mode ±1 kV differential mode Mains power quality
should be that of a
typical commercial or
hospital environment.
±2 kV common mode ±2 kV common mode
Voltage dips, short
interruptions and voltage
variations on power supply
input lines.
IEC 6
1000-4-11
<5% Ut
(>95% dip in Ut)
f
or 0.5 cycle
<12V
(>95% dip in 240V)
for 0.5 cycle
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the C-
Series Tango system
requires continued
operation during power
mains interruptions, it is
recommended that the
C-Series Tango system
be powered from an
uninterruptible power
source.
40% Ut
(60% dip in Ut)
for 5 cycles
96V
(60% dip in 240V)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cy
cles
168V
(30% dip in 24
0V)
for 25 cycles
<5% Ut
(>95%dip in Ut)
for 5 sec
<12V
(>95%dip in 24
0V)
for 5 sec
Power frequency
(50/60 Hz)
magnetic field
IEC 6
1000-4-8
3 A/m 3 A/m P
ower frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Note: Ut is the AC mains voltage prior to application of the test level.
11
English
Guidance and manufacturer’s declaration—electromagnetic immunity
The C-Series Tango system is intended for use in the electromagnetic environment specified below. The
customer or the user of the C-Series Tango system should assure that the device is used in such an environment.
Immunity test IEC60601-1-2
test level
Compliance level Electromagnetic environment—guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the C-Series Tango system, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 6
1000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms d =
1.17 √P
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
10 V/m d = 0.35 √P 80 MHz to 80
0 MHz
d = 0.70 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the C-Series Tango system
is used exceeds the applicable RF compliance level above, the C-Series Tango system should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the C-Series Tango system.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
12
Recommended separation distances between portable and mobile RF communications equipment
and the C-Series Tango system
The C-Series Tango system is intended for use in an environment in which radiated RF disturbances are
controlled. The customer or the user of the C-Series Tango system can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the C-Series Tango system as recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of transmitter (m)
Rated maximum output
power of transmitter W
150 kHz to 80 MHz
d = 1.17 √P
80 MHz to 800 MHz
d = 0.35 √P
800 MHz to 2.5 GHz
d = 0.7 √P
0.01 0.17 0.04 0.07
0.1 0.37 0.1
1 0.22
1 1.17 0.35 0.7
10 3.69 1.11 2.2
1
100 11.70 3.50 7.0
For transmit
ters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
13
Español
CPAP Tango™ C-Series
Guía Del usuario
Uso indicado
El sistema Tango CPAP es indicado para el tratamiento de la apnea obstructiva del sueño
(AOS) en pacientes que pesen más de 30 kg (66 lb). Está diseñado para uso hospitalario y
domiciliario.
PRECAUCIÓN
La ley federal estipula que este aparato solamente puede ser vendido por un médico
o por orden de éste.
Contraindicación
El tratamiento con presión positiva en las vías respiratorias puede estar contraindicado en
algunos pacientes que presenten las siguientes afecciones previas:
enfermedad bullosa pulmonar grave
neumotórax
presión arterial patológicamente baja
deshidratación
fuga de líquido cefalorraquídeo, cirugía o traumatismo craneanos recientes.
Efectos secundarios
Los pacientes deben informar al médico tratante acerca de todo dolor torácico inusual,
un dolor de cabeza severo o un aumento en su dificultad respiratoria. Una infección
respiratoria alta aguda puede requerir la suspensión temporal del tratamiento.
Los siguientes efectos secundarios pueden surgir durante el curso del tratamiento con el
Tango:
sequedad de la nariz, boca o garganta
epistaxis
hinchazón por tragar aire
molestias sinusales o del oído
irritación ocular
erupciones cutáneas.
14
Piezas del Tango
El Tango consta de:
Generador de aire Tango C-Series (aparece a continuación)
Cable de alimentación
Tubo de aire de 2 m (6’6”).
Enchufe de cable de señal
(sólo para uso con humidificador)
Salida de aire
Enchufe de cable de
alimentación
Entrada de aire
(contiene filtro de aire)
Mascarillas
Los sistemas de mascarillas ResMed recomendados para ser utilizados con el Tango son:
Mascarilla facial Mirage™ Serie II Mascarilla nasal Meridian™
Mascarilla facial Ultra Mirage™ Mascarilla nasal Mirage™
Sistema de almohadillas nasales Mirage Swift™ Mascarilla nasal Mirage Activa™
Sistema de almohadillas nasales Mirage Swift II Mascarilla nasal Mirage Vista™
Mascarilla nasal Papillon™ Mascarilla nasal Ultra Mirage™ II
Mascarilla nasal Silent Papillon™ Mascarilla facial Mirage Liberty™
Mascarilla facial Mirage Quattro™
Para información sobre el uso de las mascarillas, consulte el manual de su mascarilla.
Humidificadores
Para obtener información acerca del uso del humidificador térmico C-Series de Resmed o
del ResMed Passover, consulte el manual del humidificador.
15
Español
La interfaz del Tango
Indicador de calefacción
Indicador de
alimentación
Modo de espera Cuando se enciende el Tango, se enciende el indicador de alimentación y las horas de
funcionamiento aparecen en el visor. Todos los cambios a las configuraciones de altitud y
rampa se hacen en el modo de espera.
Rampa Configurar un tiempo de rampa permite que el tratamiento comience con una presión
baja y aumente hasta llegar a presión completa durante el tiempo de rampa (consulte
“Configurar tiempo de rampa”).
Altitud Los cambios de altitud inciden en la presión suministrada por el Tango. La función
de altitud compensa los cambios de presión relativos a cambios de altitud (consulte
“Configurar altitud”).
Horas de
funcionamiento
Las horas de funcionamiento aparecen cuando el Tango está en modo de espera.
Humidificador Los botones del humidificador se utilizan cuando se usa el humidificador térmico
C-Series y al configurar la altitud. El indicador de calefacción se ilumina cuando está
encendido el humidificador.
Instalación del Tango
PRECAUCIÓN
Tenga cuidado de no colocar el equipo en un lugar donde pueda ser golpeado o
donde alguien pueda tropezar con el cable de alimentación.
Si coloca el equipo en el piso, asegúrese de que el área esté libre de polvo y que
no haya ropa, ropa de cama u otros objetos que pudieran obstruir la entrada de
aire.
Asegúrese de que el área alrededor del equipo esté limpia y seca.
ADVERTENCIA
Asegúrese de que el cable de alimentación y el enchufe estén en buenas
condiciones y que el equipo no esté dañado.
Sólo debe usarse un tubo de aire ResMed con el equipo. Un tubo de aire
diferente podría modificar la presión que de hecho recibe, lo que reduciría la
eficacia del tratamiento.
Bloquear la manguera y/o la entrada de aire del dispositivo mientras está en
operación podría producir un sobrecalentamiento del dispositivo.
16
1
3
4
2
Conecte el cable de alimentación a la parte trasera del Tango.
Enchufe el otro extremo a una toma de corriente.
Conecte el tubo de aire a la salida de aire del Tango.
Conecte el sistema de mascarilla armado al extremo libre del tubo de aire.
Uso del Tango
Inicio del
tratamiento
Presione
cuando el Tango esté en modo de espera. Aparece un símbolo de guión
circulante en el visor.
Si está configurada la rampa, aparece en el visor el tiempo de rampa. Cuando concluye la
rampa, el tratamiento sigue con la presión prescrita.
Si no se ha configurado la rampa, el tratamiento comienza de inmediato con la presión
prescrita.
Detención del
tratamiento
Presione
.
Configurar
el tiempo de
rampa
Presione
varias veces hasta que en el visor aparezca el tiempo que desee.
Presione
para aceptar los cambios y comenzar el tratamiento. (El Tango vuelve al
modo de espera después de 5 segundos si no se presiona ningún botón.)
Las configuraciones de rampa son (en minutos): Apag, 10, 20, 45.
1.
2.
Configurar
altitud
Mantenga
presionado durante 3 segundos hasta que aparezca la configuración
de altitud actual.
Para cambiar la configuración de altitud, presione
o .
Presione
para aceptar los cambios y volver al modo de espera. (El Tango vuelve al
modo de espera después de 10 segundos si no se presiona ningún botón.)
Las configuraciones de altitud son:
1 (0–609 m/0–2000 pies)
2 (610–1219 m/2001
–4000 pies)
3 (1220–1828 m/4001–6000 pies)
4 (1829–2591 m/6001–8500 pies).
1.
2.
3.
1.
2.
3.
4.
17
Español
Limpieza y mantenimiento
ADVERTENCIA
Peligro de electrocución. No sumerja en agua el generador de aire ni tampoco el
cable de alimentación. Desconecte siempre el generador de aire antes de limpiarlo y
asegúrese de que esté seco antes de volver a conectarlo.
PRECAUCIÓN
No cuelgue el tubo de aire a la luz solar directa, ya que puede endurecerse y
agrietarse con el tiempo.
No utilice blanqueadores, cloro, alcohol, soluciones aromáticas, jabones
humectantes ni antibacterianos ni aceites aromatizados para limpiar el tubo de
aire del Tango. Estas soluciones podrían endurecer el producto y reducir la vida
útil del mismo.
Diariamente: Desconecte el tubo de aire y cuélguelo en un lugar limpio y seco hasta el
próximo uso.
Semanalmente:
Saque el tubo de aire y lávelo con agua tibia y detergente suave. Enjuáguelo bien,
cuélguelo y déjelo secar.
Antes de usarlo otra vez, vuelva a armar la mascarilla y el arnés para la cabeza.
Vuelva a conectar el tubo de aire.
Mensualmente
Limpie el exterior del Tango con un paño húmedo.
Verifique que el filtro de aire no tenga orificios ni esté obstruido por suciedad.
Reemplácelo cada seis meses, o con mayor frecuencia si se usa en un entorno con
mucho polvo.
Para más información sobre la limpieza de la mascarilla o del humidificador, consulte el
manual suministrado con la mascarilla o con el humidificador.
1.
2.
3.
18
Cambio del filtro de aire
Entrada de aire
Filtro de aire
PRECAUCIÓN
No lave el filtro de aire.
Con el cable de alimentación y el tubo de aire desconectados, voltee el Tango de
manera que la entrada de aire se oriente hacia usted. El filtro de aire está dentro de la
entrada de aire.
Saque el filtro de aire usado de la entrada de aire.
Coloque el filtro de aire nuevo en la entrada de aire en la dirección que indica la flecha,
con el lado azul orientado hacia usted.
Solución de problemas
Problema Causa posible Solución
El Tango no se pone en
marcha
El cable de alimentación no
está debidamente conectado.
Revise el cable de alimentación.
La toma de corriente está
defectuosa.
Pruebe otra toma de corriente.
No hay flujo de aire o
éste es muy bajo
El tubo de aire no está
debidamente conectado.
Ajuste la conexión del tubo de aire.
El tubo de aire está obstruido,
torcido o roto.
Desobstruya o despeje el tubo de aire.
Verifique que el tubo de aire no esté
perforado.
La presión de tratamiento
parece ser demasiado
baja o demasiado alta
Configuración de altitud
incorrecta.
Ajuste la configuración de altitud.
El visor muestra un
mensaje de error y el
tratamiento se detiene
o no puede ponerse en
marcha
Problema con el motor. Póngase en contacto con su proveedor.
1.
2.
3.
19
Español
Especificaciones técnicas
Rango de presión de
funcionamiento:
4 a 20 cm H
2
O
Dimensiones (largo x ancho
x altura):
215 mm x 189 mm x 114 mm
(8,5 pulgadas x 7,
4 pulgadas x 4,5 pulgadas)
Peso 1,1 kg (2,4 lb)
Alimentación Rango de entrada
110 V, 60 Hz, 20 VA/
10 W (consumo de energía típico),
36 VA/15 W (máximo consumo de energía)
Temperatura de
funcionamiento
+5°C a +40°C (+41°F a +104°F)
Humedad de funcionamiento 10–95% sin condensación
T
emperatura de
almacenamiento y
transporte:
-20°C a +60°C (-
4°F a +140°F)
Humedad de
almacenamiento y
transporte:
10–95% sin condensación
Rango de presión
atmosférica
1060 hP
a (a niv
el del mar) a 680 hPa
(2591 m/8500’)
Compatibilidad
electromagnética
El producto cumple con todos los requisitos pertinentes en cuanto
a compatibilidad electromagnética (EMC por sus siglas en inglés) de
acuerdo con la CEI 60601-1-2, para entornos residenciales, comerciales
y de industria ligera.
Filtro de aire Dos capas de fibra de poliéster no entrelazada, con unión de polv
o
Tubo de aire Plástico flexible, 2 m (6 pies y 6 pulgadas)
Salida de aire La salida de aire cónica de 22 mm (0,87 pulgadas) cumple con la norma
EN
1281-
1
Clasificación CEI 60601-1 Clase II (doble aislamiento), Tipo CF
El fabricante se reserva el derecho de cambiar estas especificaciones sin previo aviso.
Símbolos que aparecen en el Tango
Atención. Consulte los documentos adjuntos; Equipo tipo CF; Equipo de clase II;
IPX1
A prueba de goteo; Asociación canadiense de estándares.
20
Advertencias y precauciones generales
Una advertencia alerta sobre la posibilidad de que se produzcan lesiones.
Lea por completo la guía antes de usar el Tango.
Las indicaciones de esta guía no deben suplantar las instrucciones entregadas por el
médico tratante.
El Tango sólo debe usarse con mascarillas (y conectores*) recomendadas por
ResMed, o por un médico o terapeuta respiratorio. No debe usarse una mascarilla a
menos que el Tango esté encendido y funcionando correctamente. Nunca se debe
obstruir el o los orificios de ventilación asociados con la mascarilla.
Explicación: El Tango está diseñado para ser usado con mascarillas (o conectores
*
)
especiales con orificios de ventilación que permiten un flujo continuo de aire desde la
mascarilla.
Mientras el equipo está encendido y funcionando correctamente, el aire fresco del
dispositivo desplaza el aire espirado hacia fuera de la mascarilla a través de sus
orificios de ventilación. Si el dispositivo no está funcionando, no se proporciona
suficiente aire fresco por la mascarilla, y es posible que se vuelva a respirar aire
exhalado. Volver a respirar aire exhalado durante varios minutos puede en ocasiones
conducir a la asfixia. Si se produce una falla eléctrica o si falla el dispositivo, retire la
mascarilla.
El Tango puede configurarse para suministrar presiones de hasta 20 cm H
2
O. En el
caso poco probable de ciertas situaciones de fallas, es posible que se suministren
presiones de hasta 30 cm H
2
O.
Si se utiliza oxígeno con este equipo, el flujo del mismo deberá estar cerrado mientras
el equipo no esté en funcionamiento. Si se deja activado el oxígeno mientras el
dispositivo no está funcionando, desconecte el oxígeno y espere 30 minutos antes de
volver a encender el generador de aire.
Explicación: mientras el equipo CPAP no esté funcionando, si se deja abierto el flujo
de oxígeno, el oxígeno suministrado al tubo de respiración podrá acumularse dentro
del dispositivo CPAP, creando así un riesgo de incendio.
El oxígeno favorece la combustión. No utilice oxígeno mientras esté fumando ni en
presencia de llamas expuestas.
Asegúrese siempre de que esté fluyendo aire desde el dispositivo antes de encender
el suministro de oxígeno.
Siempre apague el suministro de oxígeno antes de detener el flujo de aire del equipo.
Nota: A un determinado caudal de oxígeno complementario, la concentración de
oxígeno inhalado variará según dónde se introduzca el oxígeno, los parámetros de
presión, el ritmo respiratorio del paciente, la selección de la mascarilla y el índice de
fuga.
P
eligro de explosión: no utilizar cerca de anestésicos inflamables.
No utilice el Tango si presenta def
ectos externos obvios o si ocurren cambios
inexplicados en su funcionamiento.
No abra la caja del Tango. No hay piezas dentro que el usuario pueda reparar. Las
reparaciones y el mantenimiento interno deberán ser efectuados únicamente por un
agente de servicio autorizado.
* Pueden integrarse puertos en la mascarilla o en los conectores próximos a la mascarilla.
21
Español
Una precaución explica medidas especiales para el uso seguro y eficaz del dispositivo.
A bajas presiones, el flujo a través de los puertos de exhalación de la mascarilla
puede no eliminar todos los gases exhalados del tubo. Puede darse cierto grado de
reinspiración.
El flujo de aire generado por este dispositivo para la respiración puede llegar a
estar hasta 6°C (11°F) por encima de la temperatura ambiente. T
enga cuidado si la
temperatura ambiente es de más de 32°C (90°F).
Servicio
El Tango debe ser inspeccionado por un centro de servicio autorizado de ResMed 5 años
después de la fecha de fabricación. Antes de esto, el dispositivo está diseñado para
proporcionar un funcionamiento seguro y confiable si se opera y mantiene de acuerdo
con las instrucciones provistas por ResMed. Los detalles acerca de la garantía acompañan
el dispositivo al momento de la entrega original de éste. Al igual que con todos los
dispositivos eléctricos, si se hace evidente la presencia de una irregularidad, debe hacer
revisar el dispositivo por un centro de servicio autorizado de ResMed.
Garantía limitada
ResMed garantiza que su producto ResMed estará libre de todo defecto de material y
mano de obra durante el plazo estipulado a continuación a partir de la fecha de adquisición
por parte del consumidor inicial. Esta garantía no es transferible.
Producto Plazo de la
garantía
Humidificadores ResMed, ResControl™, ResLink™,
ResTraxx™
1 año
Generadores de aire ResMed 2 años
Accesorios, cámaras de agua y sistemas de mascarillas
(incluido el armazón de mascarilla, la almohadilla, el
arnés para la cabeza y el tubo). Excluy
e los dispositivos
desechables.
90 días
Nota: Algunos modelos no se encuentran disponibles en todas las regiones.
Si el producto falla bajo condiciones de utilización normales, ResMed reparará o
reemplazará, a opción suya, el producto defectuoso o cualquiera de sus componentes.
Esta garantía limitada no cubre:
a) cualquier daño causado como resultado de una utilización indebida, abuso, modificación
o alteración del producto;
b) reparaciones llevadas a cabo por cualquier organización de servicio que no haya sido
expresamente autorizada por ResMed para efectuar dichas reparaciones;
c) cualquier daño o contaminación debido a humo de cigarrillo, pipa, cigarro u otras
fuentes de humo;
d) cualquier daño causado por el derrame de agua sobre el generador de aire o hacia su
interior.
La garantía queda anulada si el producto se vende o revende fuera de su región de
compra original. Las reclamaciones de garantía con respecto a productos defectuosos
deben ser realizadas por el consumidor original en el punto de compra.
22
Esta garantía revoca cualquier otra garantía expresa o implícita, incluso cualquier garantía
implícita de comerciabilidad o idoneidad para un propósito en particular. Algunas regiones
o estados no permiten limitaciones respecto a la duración de una garantía implícita, por lo
que es posible que lo que precede no sea aplicable en su caso.
ResMed no será responsable de ningún daño incidental ni emergente que se reclame por
haber ocurrido como resultado de la venta, instalación o utilización de cualquier producto
de ResMed. Algunas regiones o estados no permiten la exclusión ni limitación de daños
incidentales o emergentes, por lo que es posible que lo que precede no sea aplicable
en su caso. Esta garantía le otorga derechos legales específicos, y es posible que usted
tenga otros derechos que pueden variar de una región a otra.
Para más información sobre los derechos que le otorga esta garantía, póngase en
contacto con su distribuidor local de ResMed o con una oficina de ResMed.
Guía y declaración del fabricante – Emisiones e inmunidad electromagnéticas
Guía y declaración del fabricante – emisiones electromagnéticas
El sistema Tango C-Series está diseñado para ser utilizado en el entorno electromagnético especificado a
continuación. El cliente o usuario del sistema Tango C-Series deberá asegurarse de que sea utilizado en dicho
entorno.
Prueba de emisiones Cumplimiento Entorno electromagnético – guía
Emisiones de RF CISPR11 Grupo 1 El sistema Tango C-Series usa energía de RF sólo para su
funcionamiento interno. Por lo t
anto, sus emisiones de RF son
muy bajas y es improbable que provoquen interferencias en
equipos electrónicos cercanos.
Emisiones de RF CISPR11 Clase B El sistema Tango C-Series es apropiado para ser usado en
todos los ámbitos, incluidos ámbitos domésticos y aquéllos
conectados directamente a la red pública de bajo volt
aje que
abastece a los edificios destinados a vivienda.
Emisiones armónicas
CEI 61000-3-2
Clase A
Fluctuaciones de voltaje /
Emisiones de parpadeo
(flic
ker) CEI 61000-3-3
Cumple
Un equipo médico eléctrico requiere precauciones especiales en lo concerniente a la compatibilidad
electromagnética, y debe ser instalado y puesto en funcionamiento según la información relativ
a a dicha
compatibilidad que se proporciona en este documento.
Advertencias: El sistema Tango C-Series no debe utilizarse al lado de otro equipo ni montado sobre o debajo de él.
Si dicha proximidad o el tenerlo montado sobre o debajo de otro equipo fueran inevitables, el sistema Tango
C-Series debería ser observado para verificar su normal funcionamiento en la configuración en la que será
utilizado.
No se recomienda el uso de accesorios (por ej.: humidificadores) distintos a los que se especifican en este
manual. Dichos humidificadores podrían provocar un aumento de las emisiones o una disminución de la
inmunidad del sistema Tango C-Series.
23
Español
Guía y declaración del fabricante – inmunidad electromagnética
El sistema Tango C-Series está diseñado para ser utilizado en el entorno electromagnético especificado a
continuación. El cliente o usuario del sistema Tango C-Series deberá asegurarse de que sea utilizado en dicho
entorno.
Prueba de inmunidad Nivel de prueba
CEI 60601-1-2
Nivel de cumplimiento Entorno
electromagnético
– guía
Descarga electrostática
CEI 6
1000-4-2
± 6 kV por contacto
± 8 kV al aire
± 6 kV por contacto
± 8 kV al aire
L
os pisos deben ser
de madera, cemento o
baldosas de cerámica.
Si los pisos están
revestidos de material
sintético, la humedad
relativa debería ser al
menos 30%.
Transitorios eléctricos
rápidos/en ráfagas
CEI 61000-4-4
± 2 kV para líneas de
suministro de energía
± 2 kV La calidad de la red de
energía eléctrica deberá
ser la de un típico
entorno comercial u
hospitalario.
± 1
kV para líneas de
entrada y de salida
No corresponde
Subida rápida de tensión
CEI 61000-4-5
± 1 kV en modo dif
erencial ± 1 kV en modo diferencial La calidad de la red de
energía eléctrica deberá
ser la de un típico
entorno comercial u
hospitalario.
± 2 kV en modo común ± 2 kV en modo común
Caídas de voltaje,
variaciones de voltaje e
interrupciones cortas en
las líneas de entrada del
suministro de energía
CEI 6
1000-4-11
<5% Ut
(>95% de caída con
respecto al Ut)
durante 0,5 ciclos
<
12V
(>95% de caída en 240 V)
durante 0,5 ciclos
La calidad de la red de
energía eléctrica deberá
ser la de un típico
entorno comercial u
hospitalario.
Si el usuario del
sistema Tango
C-Series necesita
un funcionamiento
continuo a pesar de
cortes en el suministro
de energía, se
recomienda que el
sistema Tango C-Series
sea alimentado por
una fuente de energía
continua.
40% del Ut
(60% de caída con
respecto al Ut)
durante 5 ciclos
96 V
(60% de caída en 240 V)
durante 5 ciclos
70% del Ut
(30% de caída con
respecto al Ut)
durante 25 ciclos
168 V
(30% de caída en 240 V)
durante 25 ciclos
<5% Ut
(>95% de caída con
respecto al Ut)
durante 5 seg
<
12 V
(>95% de caída en 240
V)
durante 5 seg
Campo magnético
de la frecuencia industrial
(50/60 Hz)
CEI 6100
0-4-8
3 A/m 3 A/m L
os campos
magnéticos de la
frecuencia industrial
deben tener los niveles
propios de un local
típico de un entorno
comercial u hospitalario
típico.
Nota: Ut es la tensión de red de CA antes de la aplicación del nivel de prueba.
24
Guía y declaración del fabricante – inmunidad electromagnética
El sistema Tango C-Series está diseñado para ser utilizado en el entorno electromagnético especificado a
continuación. El cliente o usuario del sistema Tango C-Series deberá asegurarse de que sea utilizado en dicho
entorno.
Prueba de
inmunidad
CEI60601-1-2
Nivel de prueba
Nivel de
cumplimiento
Entorno electromagnético – guía
Los equipos de comunicaciones por RF
portátiles y móviles no deben ser usados
cerca de parte alguna del sistema Tango
C-Series (incluidos los cables) a una distancia
menor que la recomendada, calculada a partir
de la ecuación aplicable a la frecuencia del
transmisor.
Distancia recomendada
RF conducida
CEI 6
1000-4-6
3 Vrms
de 150 kHz a 80 MHz
3
Vrms d = 1,17 √P
RF irradiada
CEI 61000-4-3
10 V/m
de 80 MHz a 2,5 GHz
10 V/m d = 0,35 √P 80 MHz a 80
0 MHz
d = 0,70 √P 800 MHz a 2,5 GHz
Donde P es la potencia nominal máxima de
salida del transmisor en vatios (W) según el
fabricante del transmisor y d es la distancia
recomendada en metros (m).
Las intensidades de campo de transmisores
de RF fijos, determinadas según un control
electromagnético del lugar,
a
deben ser
menores que el nivel de cumplimiento en
cada rango de frecuencias.
b
Puede haber interferencias cerca de equipos
marcados con el siguiente símbolo:
Nota 1: A 80 MHz y 800 MHz se aplica el rango de frecuencias más alto.
Nota 2: Es posible que estas guías no se apliquen en todas las situaciones. La propagación electromagnética es
afectada por la absorción y el reflejo en estructuras, objetos y personas.
a
Las intensidades de campo de los transmisores fijos, tales como estaciones base para teléfonos de radio
(celulares/inalámbricos) y radios móviles terrestres, radio amateur, transmisores de radio AM y FM y de
televisión, no se pueden predecir con exactitud en forma teórica. Para evaluar el entorno electromagnético
provocado por transmisores de RF fijos, se debería considerar la realización de un control electromagnético del
lugar. Si la intensidad de campo medida en el lugar donde habrá de usarse el sistema Tango C-Series excediera
el nivel de cumplimiento de RF correspondiente indicado antes, el sistema Tango C-Series debería ser vigilado
en cuanto a su normal funcionamiento. Si ocurriera un funcionamiento anormal, pueden ser necesarias medidas
adicionales, como la reorientación o reubicación del sistema Tango C-Series.
b
En el rango de frecuencias de 150 kHz a 80 MHz, las intensidades de campo deberían ser menores de 10 V/m.
25
Español
Distancias recomendadas entre equipos de comunicación por RF portátiles y móviles y el sistema
Tango C-Series
El sistema Tango C-Series está diseñado para ser usado en un entorno en el que las alteraciones por RF irradiada
sean controladas. El cliente o usuario del sistema Tango C-Series puede ayudar a prevenir la interferencia
electromagnética guardando una distancia mínima entre los equipos de comunicaciones por RF portátiles y
móviles (transmisores) y el sistema Tango C-Series, tal como se recomienda a continuación según la potencia
máxima de salida del equipo de comunicaciones.
Distancia según la frecuencia del transmisor (m)
Potencia nominal máxima
de salida del transmisor W
150 kHz a 80 MHz
d = 1,17 √P
80 MHz a 800 MHz
d = 0,35 √P
800 MHz a 2,5 GHz
d = 0,7 √P
0,01 0,17 0,04 0,07
0,1 0,37 0,
11 0,22
1 1,17 0,35 0,7
10 3,69 1,1
1 2,21
100 11,70 3,50 7
,0
Para transmisores con una potencia nominal máxima de salida que no figure en la tabla precedente, la distancia
recomendada d en metros (m) se puede determinar por medio de la ecuación que se aplica a la frecuencia del
transmisor, donde P es la potencia nominal máxima de salida del transmisor en vatios (W), según el fabricante del
transmisor.
Nota 1: A 80 MHz y 800 MHz, se aplica la distancia para el rango de frecuencias más alto.
Nota 2: Es posible que estas guías no se apliquen en todas las situaciones. La propagación electromagnética es
afectada por la absorción y el reflejo en estructuras, objetos y personas.
26
27
Français
C-Series Tango™
GuiDe utilisateur
Usage prévu
L’appareil de PPC Tango est prévu pour le traitement du syndrome d’apnée obstructive du
sommeil (SAOS) chez les patients de plus de 66 lb (30 kg). Il est prévu pour une utilisation
à domicile et en milieu médical.
ATTENTION
Selon la loi fédérale américaine, ce dispositif ne peut être vendu aux U.S.A. que par
un médecin ou sur prescription médicale.
Contre-indications
Le traitement par PPC est contre-indiqué chez certains patients présentant l’une des
pathologies préexistantes suivantes :
affection pulmonaire bulleuse grave
pneumothorax
hypotension
déshydratation
fuite de liquide céphalo-rachidien, récent traumatisme crânien ou intervention
chirurgicale crânienne.
Effets secondaires
Les patients doivent avertir leur médecin traitant en cas de douleurs thoraciques
inhabituelles, de maux de tête sévères ou d’une dyspnée accrue. Une infection aiguë des
voies aériennes supérieures peut nécessiter l’arrêt temporaire du traitement.
Les effets secondaires suivants peuvent apparaître au cours du traitement sous T
ango :
sécheresse nasale, buccale ou de la gorge
saignements de nez
ballonnements
gêne au niveau de l’oreille ou des sinus
irritation des yeux
érythèmes cutanés.
28
Composants du Tango
Votre Tango se compose des éléments suivants :
Appareil C-Series Tango (représenté ci-dessous)
Câble d’alimentation
Circuit respiratoire 6’6” (2 m).
Prise du câble d’interface
(avec un humidificateur
uniquement)
Sortie d’air
Prise du câble
d’alimentation
Entrée d’air
(contient le filtre à air)
Masques
Les masques ResMed suivants sont recommandés pour utilisation avec le Tango :
Masque facial Mirage™ Série 2 Masque nasal Meridian™
Masque facial Ultra Mirage™ Masque nasal Mirage™
Masque narinaire Mirage Swift™ Masque nasal Mirage Activa™
Masque narinaire Mirage Swift II Masque nasal Mirage Vista™
Masque nasal Papillon™ Masque nasal Ultra Mirage™ II
Masque nasal Silent Papillon™ Masque facial Mirage Liberty™
Masque facial Mirage Quattro™
Pour les informations concernant l’utilisation des masques, veuillez vous référer au
manuel de votre masque.
Humidificateurs
Pour les informations concernant l’utilisation de l’humidificateur chauffant C-Series
ResMed ou Passover ResMed, veuillez vous référer au manuel de votre humidificateur.
29
Français
Interface du Tango
Témoin de chauffe
Témoin
d’alimentation
Mode d’attente Lorsque le Tango est sous tension, le témoin d’alimentation est allumé et les
heures de fonctionnement sont affichées sur l’écran LCD. Toutes les modifications
des réglages d’altitude et de rampe doivent être effectuées en mode d’attente.
Rampe Le réglage d’une durée de rampe permet de démarrer le traitement à une pression
basse, qui augmente progressivement jusqu’à la pression de traitement sur la
durée de rampe (voir « Réglage de la durée de rampe »).
Altitude L
e changement d’altitude a une incidence sur la pression délivrée par le Tango. La
fonction d’altitude permet de compenser les changements de pression liés aux
changements d’altitude (voir « Réglage de l’altitude
»).
Heures de
fonctionnement
Les heures de fonctionnement sont affichées lorsque le Tango est en mode
d’attente.
Humidificateur Les touches de commande de l’humidificateur sont utilisées lorsque
l’humidificateur chauffant C-Series est utilisé et lors du réglage de l’altitude. Le
témoin de chauffe est allumé lorsque l’humidificateur est en marche.
30
Installation du Tango
ATTENTION
Veiller à placer l’appareil de telle manière que personne ne puisse le faire tomber
ou se prendre les pieds dans le câble d’alimentation.
Si l’appareil est posé sur le sol, veiller à ce qu’il le soit dans un endroit exempt de
poussière, de literie, de vêtements ou de tout autre objet susceptible d’obstruer
l’entrée d’air.
Vérifier que la zone autour de l’appareil est propre et sèche.
AVERTISSEMENT
Vérifier que le câble d’alimentation et la fiche sont en bon état et que
l’équipement n’est pas endommagé.
Seul le circuit respiratoire ResMed peut être utilisé avec l’appareil. Les circuits
respiratoires d’un autre type risquent de modifier la pression reçue et de réduire
l’efficacité du traitement.
L’obstruction du circuit ou de l’entrée d’air pendant le fonctionnement risque de
causer une surchauffe de l’appareil.
1
3
4
2
Branchez le câble d’alimentation à l’arrière du Tango.
Branchez l’autre extrémité du câble d’alimentation à une prise de courant.
Raccordez le circuit respiratoire à la sortie d’air du Tango.
Raccordez le masque assemblé à l’extrémité libre du circuit respiratoire.
1.
2.
3.
4.
31
Français
Utilisation du Tango
Démarrage du
traitement
Appuyez sur la touche
lorsque le Tango est en mode d’attente. Un cercle en pointillés
est affiché sur l’écran LCD.
Si la rampe est réglée, sa durée est affichée sur l’écran LCD. Une fois qu’elle s’est
écoulée, le traitement se poursuit à la pression prescrite.
Si aucune rampe n’est réglée, le traitement démarre immédiatement à la pression
prescrite.
Arrêt du
traitement
Appuyez sur la touche
.
Réglage de
la durée de
rampe
Appuyez plusieurs fois sur la touche
jusqu’à ce que le réglage voulu s’affiche sur
l’écran LCD.
Appuyez sur la touche
pour accepter les modifications de réglage et démarrer
le traitement. (Le Tango retourne en mode d’attente au bout de 5 secondes si vous
n’appuyez sur aucune autre touche.)
Les options de réglage de rampe sont les suivantes (en minutes) : Off, 10, 20,
45.
1.
2.
Réglage de
l’altitude
Maintenez la touche
enfoncée pendant 3 secondes jusqu’à ce que le réglage
d’altitude actuel s’affiche.
Appuyez sur la touche
ou pour modifier le réglage d’altitude.
Appuyez sur la touche
pour accepter les modifications de réglage et retourner en
mode d’attente. (Le Tango retourne en mode d’attente au bout de 10 secondes si
vous n’appuyez sur aucune autre touche.)
Les options de réglage pour l’altitude sont les suivantes :
1 (0 – 2000’ / 0 – 609 m),
2 (2001 – 40
00’ / 610 – 1219 m),
3 (4001 – 6000’ / 1220 – 1828 m),
4 (6001 – 8500’ / 1829 – 2591 m).
1.
2.
3.
Nettoyage et entretien
AVERTISSEMENT
Risque de chocs électriques. Ne pas immerger l’appareil ni le câble d’alimentation
dans l’eau. Systématiquement débrancher l’appareil avant de le nettoyer et vérifier
qu’il est bien sec avant de le brancher à nouveau.
ATTENTION
Ne pas suspendre le circuit respiratoire à la lumière directe du soleil afin d’éviter
qu’il ne durcisse ou ne se fissure avec le temps.
Ne pas utiliser d’eau de Javel, de chlore, d’alcool ou de solutions à base
d’aromates, ni de savons hydratants ou antibactériens ou d’huiles parfumées
pour le lavage du circuit respiratoire ou du Tango. Ces solutions peuvent causer
le durcissement du produit et réduire sa durée de vie.
Quotidien : détachez le circuit respiratoire et suspendez-le dans un endroit propre et sec
jusqu’à la prochaine utilisation.
Hebdomadaire :
Détachez le circuit respiratoire et nettoyez-le à l’eau tiède avec un détergent doux.
Rincez soigneusement et suspendez-le pour le faire sécher.
Remontez le masque et le harnais avant de les réutiliser.
Raccordez le circuit respiratoire à nouveau.
1.
2.
3.
32
Mensuel :
Nettoyez l’extérieur du Tango à l’aide d’un chiffon humide.
Vérifiez que le filtre à air n’est ni percé ni bloqué par de la saleté. Remplacez-le tous les
six mois ou plus souvent si l’appareil est utilisé dans un endroit poussiéreux.
Pour les informations concernant le nettoyage du masque ou de l’humidificateur, veuillez
vous référer au manuel fourni avec le masque ou l’humidificateur.
Remplacement du filtre à air
Entrée d’air
Filtre à air
ATTENTION
Ne pas laver le filtre à air.
Avec le câble d’alimentation débranché et le circuit respiratoire détaché, mettez le
Tango à l’envers afin que l’entrée d’air se trouve face à vous. Le filtre à air se trouve
dans l’entrée d’air.
Tirez sur le filtre à air usagé pour le retirer de l’entrée d’air.
Mettez un nouveau filtre à air en place dans l’entrée d’air dans le sens indiqué par la
èche, avec la face bleue face à vous.
Stratégie de dépannage
Problème Cause possible Solution
Le Tango ne démarre pas. Le câble d’alimentation n’est
pas correctement branché.
Vérifiez le câble d’alimentation.
La prise de courant est
défectueuse.
Essayez une autre prise de courant.
Absence de débit d’air/
Débit d’air faible.
Le circuit respiratoire n’est pas
correctement raccordé.
Ajustez le raccordement du circuit
respiratoire.
Le circuit respiratoire est
obstrué, pincé ou percé.
Débouchez ou désentortillez le circuit
respiratoire. Vérifiez que le circuit
respiratoire n’est pas percé.
La pression de traitement
semble basse ou trop
élevée.
Réglage de l’altitude incorrect. Ajustez le réglage d’altitude.
Un message d’erreur est
affiché sur l’écran LCD et
le traitement s’arrête ou
ne démarre pas du tout.
Problème de moteur. Contactez votre prestataire de santé.
1.
2.
3.
33
Français
Caractéristiques
Pression de fonctionnement 4 à 20 cm H
2
O
Dimensions (L x l x P) 8,5” x 7,4” x 4,5”
(215
mm x 189 mm x 114 mm)
Poids 2,4 lb (1,1 kg)
Alimentation électrique Courant
110 V, 60 Hz, 20 V
A / 10 W (consommation typique de courant),
36 VA / 15 W (consommation maximale de courant)
Température de
fonctionnement
+41°F à +104°F (+5°C à +40°C)
Humidité de fonctionnement 10 à 95 % sans condensation
T
empérature de stockage et
de transport
-
4°F à +140°F (-20°C à +60°C)
Humidité de stockage et de
transpor
t
10 à 95 % sans condensation
Fourchette de pressions
atmosphériques
1060
hPa (niveau de la mer) à 680 hPa
(850
0’ / 2591 m)
Compatibilité
électromagnétique
Le produit est conforme à toutes les exigences de compatibilité
électromagnétique (CEM) applicables conformément à CEI 60601-1-2
pour les environnements résidentiels, commerciaux et l’industrie
légère.
Filtre à air Deux couches de mousse à cellules ouvertes de polyester
, imprégnées
de fibres polyester
Circuit respiratoire Plastique fl
exible, longueur 6’6” (2 m)
Sortie d’air Le raccord conique 0,87” (22
mm) de sortie d’air est conforme à la
norme EN 1281-1
Classification CEI 6060
1-1 Classe II (double isolation), Type CF
Le fabricant se réserve le droit de modifier ces caractéristiques sans notification préalable.
Symboles apparaissant sur le Tango
Attention. Veuillez consulter la documentation jointe ; Équipement de type CF ;
Équipement de classe II ;
IPX1
Protection anti-gouttes ; Association Canadienne de
Normalisation.
34
Avertissements et précautions d’ordre général
Un avertissement vous met en garde contre le risque de blessure.
Lire le manuel entièrement avant d’utiliser le Tango.
Les conseils donnés dans ce manuel ne remplacent pas les instructions du médecin
traitant.
Le Tango doit être utilisé uniquement avec les masques (et raccords*) recommandés
par ResMed, un médecin ou un kinésithérapeute respiratoire. Le masque ne peut être
porté que si le Tango est sous tension et s’il fonctionne correctement. Les orifices de
ventilation du masque ne doivent jamais être obstrués.
Explication : le Tango est conçu pour être utilisé avec des masques (ou raccords
*
)
spéciaux possédant des orifices de ventilation qui permettent l’écoulement continu de
l’air hors du masque.
Lorsque l’appareil est sous tension et qu’il fonctionne correctement, de l’air propre
provenant de l’appareil expulse l’air expiré via les orifices de ventilation du masque.
Si l’appareil ne fonctionne pas, le débit d’air propre délivré dans le masque est
insuffisant et une réinhalation de l’air expiré peut se produire. La réinhalation de
l’air expiré pendant une durée supérieure à quelques minutes peut parfois entraîner
la suffocation. Le masque doit être retiré en cas de coupure de courant ou de
dysfonctionnement de l’appareil.
L
e Tango peut être réglé pour délivrer des pressions allant jusqu’à 20 cm H
2
O. Dans le
cas, peu probable, d’une panne, les pressions peuvent s’élever jusqu’à 30 cm H
2
O.
Si de l’oxygène est utilisé avec cet appareil, l’alimentation en oxygène doit être fermée
lorsque l’appareil n’est pas en mode de fonctionnement. Si l’alimentation en oxygène
a été laissée ouverte avec l’appareil arrêté, déconnecter l’alimentation en oxygène,
puis attendre 30 minutes avant de remettre l’appareil sous tension.
Explication : lorsque l’appareil de PPC n’est pas en marche et que l’alimentation
en oxygène n’est pas coupée, l’oxygène délivré dans le circuit respiratoire peut
s’accumuler dans l’appareil et créer un risque d’incendie.
L’oxygène est inflammable. Ne pas fumer ni approcher de flamme nue de l’appareil
lors de l’utilisation d’oxygène.
Systématiquement vérifier qu’un débit d’air est généré par l’appareil avant d’ouvrir
l’alimentation en oxygène.
Systématiquement couper l’alimentation en oxygène avant d’arrêter le débit d’air de
l’appareil.
Remarque : lorsque de l’oxygène additionnel est délivré à un débit fixe, la
concentration de l’oxygène inhalé varie en fonction du point d’entrée de l’oxygène,
des réglages de pression, de la respiration du patient, du masque sélectionné ainsi
que du niveau de fuite.
Danger d’explosion – Ne pas utiliser l’appareil à proximité d’anesthésiques
inflammables.
Ne pas utiliser le Tango en cas de défauts externes évidents ou si sa performance
devient aléatoire.
Ne pas ouvrir le boîtier du Tango. Il ne contient aucune pièce pouvant être entretenue
par l’utilisateur. Les réparations et l’entretien interne doivent être effectués par un
technicien agréé.
* Des ports peuvent être intégrés au masque ou à des raccords proches du masque.
35
Français
Une précaution (Attention) explique des mesures particulières permettant d’assurer une
utilisation efficace et sûre de l’appareil.
À basses pressions, le débit aux orifices de ventilation de votre masque peut s’avérer
insuffisant pour expulser tous les gaz expirés hors du circuit respiratoire. Une
réinhalation peut se produire.
La température du débit d’air produit par cet appareil peut être supérieure de 11°F
(6°C) à la température ambiante. Des précautions devront être prises lorsque la
température ambiante est supérieure à 90°F (32°C).
Réparations
Le Tango doit être inspecté par un service technique agréé ResMed 5 ans après la date
de fabrication. Avant cette date, l’appareil devrait assurer un fonctionnement fiable et
sûr s’il est utilisé et entretenu conformément aux instructions fournies par ResMed. Les
informations de garantie sont fournies avec l’appareil lors de sa livraison initiale. Comme
avec tous les appareils électriques, si vous constatez un défaut quelconque, vous devez
faire vérifier l’appareil par un service technique agréé ResMed.
Garantie limitée
ResMed garantit votre produit ResMed contre tout défaut de matériaux et de
main-d’œuvre pour la période spécifiée ci-après à compter de la date d’achat par le client
d’origine. Cette garantie n’est pas cessible.
Produit Durée de la
garantie
Humidificateurs ResMed, ResControl™, ResLink™,
ResTraxx™
1 an
Appareils de traitement ResMed 2 ans
Accessoires, réservoirs et masques (y compris entourage
rigide, bulle, harnais et circuit respiratoire). Les appareils
à usage unique sont ex
clus.
90 jours
Remarque : certains de ces modèles ne sont pas disponibles dans tous les pays.
En cas de défaillance du produit dans des conditions normales d’utilisation, ResMed, à
son entière discrétion, répare ou remplace le produit défectueux ou toute pièce. Cette
garantie limitée ne couvre pas :
a) tout dommage résultant d’une utilisation incorrecte, d’un usage abusif ou d’une
modification ou transformation opérée sur le produit
;
b) les réparations et l’entretien effectués par tout service de réparation sans l’autorisation
e
xpresse de ResMed ;
c) tout dommage ou contamination causé par de la fumée de cigarette, de pipe, de cigare
ou autre ;
d) tout dommage causé par de l’eau renversée sur ou dans un appareil.
La garantie est annulée pour les produits vendus ou revendus dans un pays autre que
celui où ils ont été achetés à l’origine. Les réclamations au titre de la garantie pour les
produits défectueux doivent être présentées au lieu d’achat par le client initial.
Cette garantie remplace toute autre garantie expresse ou implicite, y compris toute
garantie implicite de qualité marchande ou d’adéquation à un usage particulier. Certains
pays ou états n’autorisent pas les limitations de durée pour les garanties implicites ; il est
donc possible que la limit
ation susmentionnée ne s’applique pas à votre cas.
36
ResMed ne peut être tenue pour responsable de tout dommage accessoire ou indirect
résultant de la vente, de l’installation ou de l’utilisation de tout produit ResMed. Certains
pays ou états n’autorisent ni l’exclusion ni la limitation des dommages accessoires ou
indirects ; il est donc possible que la limitation susmentionnée ne s’applique pas à votre
cas. Cette garantie vous octroie des droits reconnus par la loi, et v
ous pouvez également
avoir d’autres droits qui varient en fonction du pays où vous habitez.
Pour de plus amples informations sur vos droits de garantie, veuillez contacter votre
prestataire de santé ou ResMed.
Guide et déclaration du fabricant – Émissions et immunité électromagnétiques
Guide et déclaration du fabricant – Émissions électromagnétiques
Le C-Series Tango est prévu pour une utilisation dans l’environnement électromagnétique décrit ci-dessous. Le
client ou l’utilisateur du C-Series Tango doit s’assurer que celui-ci est utilisé dans un environnement de ce type.
Essai de contrôle des
émissions
Conformité Environnement électromagnétique – Guide
Émissions RF CISPR11 Groupe 1 Le C-Series Tango utilise de l’énergie RF uniquement pour son
fonctionnement interne. P
ar conséquent, les émissions RF
sont très faibles et ne devraient pas causer d’interférences
avec l’équipement électronique environnant.
Émissions RF CISPR 11 Classe B Le C-Series Tango peut être utilisé dans tous les
établissements, y compris les ét
ablissements domestiques et
ceux directement raccordés au réseau public de distribution à
basse tension qui fournit de l’électricité aux bâtiments utilisés
à des fins domestiques.
Ra
yonnements
harmoniques
CEI 61000-3-2
Classe A
Émissions dues aux
fluctuations de tension/au
papillotement
CEI 6100
0-3-3
Conforme
Des précautions particulières doivent être prises avec les appareils électromédicaux quant à la compatibilité
électromagnétique, et leur inst
allation et leur mise en service doivent être effectuées conformément aux
informations sur la compatibilité électromagnétique fournies dans ce document.
Avertissements : le C-Series Tango ne doit pas être utilisé à proximité d’autres équipements ni posé sur ou sous
d’autres équipements.
S’il est impossible de l’utiliser autrement, vous devez vous assurer qu’il f
onctionne normalement lorsqu’il est
placé à l’endroit où il sera utilisé.
Il est déconseillé d’utiliser des accessoires (par exemple des humidificateurs) autres que ceux spécifiés dans ce
manuel. Ils peuvent entraîner une augmentation des émissions ou une réduction de l’immunité du
C-Series Tango.
37
Français
Guide et déclaration du fabricant – Immunité électromagnétique
Le C-Series Tango est prévu pour une utilisation dans l’environnement électromagnétique décrit ci-dessous. Le
client ou l’utilisateur du C-Series Tango doit s’assurer que celui-ci est utilisé dans un environnement de ce type.
Essai de contrôle de
l’immunité
Niveau d’essai
CEI 60601-1-2
Niveau de conformité Environnement
électromagnétique
– Guide
Décharge électrostatique
CEI 6
1000-4-2
Contact ± 6 kV
Air ± 8 kV
Contact ± 6 kV
Air ± 8 kV
L
es sols doivent être
en bois, en ciment
ou carrelés. Si le
revêtement des sols est
synthétique, l’humidité
relative doit être d’au
moins 30 %.
Transitoires/salves
électriques rapides
CEI 61000-4-
4
± 2 kV pour les lignes
d’alimentation
± 2 kV La qualité de
l’alimentation secteur
doit être celle d’un
environnement
commercial ou médical
standard.
± 1 kV pour les lignes
d’alimentation/de
signalisation
Sans objet
Surtension
CEI 61000-4-5
± 1 kV
en mode différentiel ± 1 kV en mode différentiel La qualité de
l’alimentation secteur
doit être celle d’un
environnement
commercial ou médical
standard.
± 2 kV en mode commun ± 2 kV en mode commun
Baisses de tension, brèves
coupures de courant et
variations de tension sur
les lignes d’alimentation
CEI 6
1000-4-11
< 5 % Ut
(> 95 % de baisse en Ut)
pour 0,5 cy
cle
< 12 V
(> 95 % de baisse en
240 V)
pour 0,5 cycle
La qualité de
l’alimentation secteur
doit être celle d’un
environnement
commercial ou médical
standard.
Si l’utilisateur du
C-Series Tango ne
peut pas supporter
l’interruption du
traitement en cas de
coupure de courant,
il est recommandé
d’utiliser le C-Series
Tango avec une
source d’aliment
ation
ininterruptible.
40 % Ut
(60 % de baisse en Ut)
pour 5 cycles
96 V
(60 % de baisse en 240 V)
pour 5 cycles
70 % Ut
(30 % de baisse en Ut)
pour 25 cycles
168 V
(30 % de baisse en 2
40 V)
pour 25 cycles
< 5 % Ut
(> 95 % de baisse en Ut)
pour 5 sec
< 12 V
(> 95 % de baisse en
2
40 V)
pour 5 sec
Champ magnétique
de la fréquence
d’alimentation (50/60 Hz)
CEI 610
00-4-8
3 A/m 3 A/m L
es champs
magnétiques
de la fréquence
réseau doivent
correspondre à ceux
d’un environnement
commercial ou médical
standard.
Remarque : Ut correspond à la tension secteur AC avant l’application du niveau d’essai.
38
Guide et déclaration du fabricant – Immunité électromagnétique
Le C-Series Tango est prévu pour une utilisation dans l’environnement électromagnétique décrit ci-dessous. Le
client ou l’utilisateur du C-Series Tango doit s’assurer que celui-ci est utilisé dans un environnement de ce type.
Essai de
contrôle de
l’immunité
CEI60601-1-2
Niveau d’essai
Niveau de
conformité
Environnement électromagnétique
– Guide
Les équipements de communication RF
portables et mobiles ne doivent pas être
utilisés à proximité des composants du
C-Series
Tango, y compris les câbles, à
une distance inférieure à la distance de
séparation recommandée calculée sur la
base de l’équation applicable à la fréquence
de l’émetteur.
Distance de séparation recommandée
RF transmises
par conduction
CEI 6
1000-4-6
3 Vrms
150 kHz à 80 MHz
3 Vrms d =
1,17 √P
RF transmises
par radiation
CEI 61000-4
-3
10 V/m
80 MHz à 2,5 GHz
10 V/m d = 0,35 √P 80 MHz à 80
0 MHz
d = 0,70 √P 800 MHz à 2,5 GHz
où P correspond à la puissance nominale
de sortie maximale de l’émetteur en watts
(W) indiquée par le fabricant de l’émetteur
et d correspond à la distance de séparation
recommandée en mètres (m).
L’intensité des champs d’émetteurs RF
fixes, telle qu’elle est déterminée par
l’étude electromagnétique du site
a
doit être
inférieure au niveau de conformité pour
chaque plage de fréquences.
b
Des interférences peuvent se produire à
proximité d’équipements portant le symbole
suivant :
Remarque 1 : à 80 MHz et 800 MHz, la plage de fréquences supérieure s’applique.
Remarque 2 : il est possible que ces directives ne s’appliquent pas à toutes les situations. La propagation
électromagnétique dépend de l’absorption et de la réflexion des structures, des objets et des personnes.
a
L’intensité des champs d’émetteurs fixes, tels que des stations de base pour des radiotéléphones (cellulaires/
sans fil) et des radios mobiles, la radio amateur, les radios AM et FM et la télévision, ne peut pas être prédite
théoriquement avec précision. Pour évaluer l’environnement électromagnétique dû aux émetteurs RF fixes,
une étude électromagnétique du site doit être envisagée. Si l’intensité du champ mesurée à l’endroit où le
C-Series Tango est utilisé dépasse le niveau de conformité RF applicable figurant ci-dessus, il faudra s’assurer
du bon fonctionnement du C-Series Tango dans un tel environnement. En cas d’anomalie, des mesures
supplémentaires devront peut-être être prises, par exemple changement de position ou d’endroit pour le
C-Series Tango.
b
Pour la plage de fréquences 150 kHz à 80 MHz, l’intensité des champs doit être inférieure à 10 V/m.
39
Français
Distances de séparation recommandées entre les équipements de communication RF portables et
mobiles et le C-Series Tango.
Le C-Series Tango est prévu pour une utilisation dans un environnement dans lequel les perturbations
radioélectriques par radiation sont contrôlées. Le client ou l’utilisateur du C-Series Tango peut contribuer à la
prévention des perturbations électromagnétiques en respectant les distances minimales recommandées
ci-dessous pour la séparation entre les équipements de communication RF portables et mobiles (émetteurs) et le
C-Series Tango, en fonction de la puissance maximale de sortie de l’équipement de communication.
Distance de séparation en fonction de la fréquence de l’émetteur (m)
Puissance nominale
de sortie maximale de
l’émetteur W
150 kHz à 80 MHz
d = 1,17 √P
80 MHz à 800 MHz
d = 0,35 √P
800 MHz à 2,5 GHz
d = 0,7 √P
0,01 0,17 0,04 0,07
0,1 0,37 0,
11 0,22
1 1,17 0,35 0,7
10 3,69 1,1
1 2,21
100 11,70 3,50 7
,0
Pour les émetteurs avec une puissance nominale de sortie maximale ne figurant pas dans le tableau ci-dessus, la
distance de séparation recommandée d en mètres (m) peut être déterminée à l’aide de l’équation applicable à la
fréquence de l’émetteur, où P correspond à la puissance nominale de sortie maximale de l’émetteur en watts (W)
indiquée par le fabricant de l’émetteur.
Remarque 1 : à 80 MHz et 800 MHz, la distance de séparation pour la plage de fréquences supérieure
s’applique.
Remarque 2 : il est possible que ces directives ne s’appliquent pas à toutes les situations. La propagation
électromagnétique dépend de l’absorption et de la réflexion des structures, des objets et des personnes.
40
ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA
ResMed (UK) Ltd (EU Authorized Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK
ResMed Offices Australia, Austria, Brazil, China, Finland, France, Germany, Hong Kong, India, Japan, Malaysia, Netherlands,
New Zealand, Norway, Singapore, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com for contact details).
C-Series Tango
Protected by patents: US 5199424, US 5522382, US 6705315.
Protected by design registrations: AU 307932, AU 307933, AU 307934, AU 307935, AU 309812, AU 310045, AU 310046,
AU 310165, AU 311258, AU 311273, AU 311486, AU 311487, CN 200630002263.2, CN 20063000
2264.7, CN 200630002265.1,
CN 200630002266.6, CN 200630002267.0, CN 200630002269.X, CN 200630002270.2, CN 200630002271.7, CN 200630137884.1,
EU 599634, EU 600176, JP 1293705, JP 1294379, JP 1294380, JP 1296487, JP 1296603, JP 1296604, JP 1296605, JP 1303495,
NZ 407118, NZ 408195, NZ 408705, NZ 408706, NZ 408707, NZ 408708, NZ 408709, NZ 40871
0, US D542900, US D544591, US
D545954, US D549321, US D549810, US D550349. Patents and designs pending.
Tango, Mirage, Mirage Activa, Mirage Vista, Ultra Mirage, Mirage Swift, Mirage Liberty, Mirage Quattro and Meridian are
trademarks of ResMed Ltd. Mirage, Mirage Activa, Mirage Vista, Ultra Mirage, Mirage Swift and Meridian are registered in US
Patent and Trademark Office. Papillon and Silent Papillon are trademarks of MAP Medizin-Technologie GmbH.
© 2007 ResMed Ltd.
Global leaders in sleep and respiratory medicine www.resmed.com
34835/1 07 10
C-Series Tango
USER
AMER ML

Transcripción de documentos

C-Series Tango™ positive airway pressure Device User Guide English • Español • Français User Guide Intended Use The Tango CPAP system is indicated for the treatment of obstructive sleep apnea (OSA) in patients weighing more than 66 lb (30 kg). It is intended for home and hospital use. CAUTION Federal law restricts this device to sale by or on the order of a physician. Contraindication Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions: • severe bullous lung disease • pneumothorax • pathologically low blood pressure • dehydration • cerebrospinal fluid leak, recent cranial surgery or trauma. Adverse Effects Patients should report unusual chest pain, severe headache or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment. The following side effects may arise during the course of therapy with the Tango: • drying of the nose, mouth or throat • nosebleed • bloating • ear or sinus discomfort • eye irritation • skin rashes. Parts of the Tango Your Tango comprises: • C-Series Tango flow generator (shown below) • Power cord • 6’6” (2 m) air tubing. Signal cable socket (humidifier use only) Power cord socket Air outlet Air inlet (contains air filter)  English C-Series Tango™ Masks The ResMed mask systems recommended for use with the Tango are: • Mirage™ Full Face Mask Series II • Meridian™ Nasal Mask • Ultra Mirage™ Full Face Mask • Mirage™ Nasal Mask • Mirage Swift™ Nasal Pillows System • Mirage Activa™ Nasal Mask • Mirage Swift II Nasal Pillows System • Mirage Vista™ Nasal Mask • Papillon™ Nasal Mask • Ultra Mirage™ II Nasal Mask • Silent Papillon™ Nasal Mask • Mirage Liberty™ Full Face Mask • Mirage Quattro™ Full Face Mask For information on using masks, see your mask manual. Humidifiers For information on using the ResMed C-Series Heated Humidifier or the ResMed Passover, see your humidifier manual. The Tango Interface Heating indicator Power indicator Standby mode When the Tango is powered, the power indicator is lit and the run hours are displayed on the LCD. All changes to altitude and ramp settings are made in standby mode. Ramp Setting a ramp time enables therapy to start at a low pressure and build up to full pressure during the ramp time (see “Set ramp time”). Altitude Changes in altitude affect the pressure delivered by the Tango. The altitude feature compensates for pressure changes related to changes in altitude (see “Set altitude”). Run hours Run hours are displayed when the Tango is in standby mode. Humidifier The humidifier buttons are used when using the C-Series Heated Humidifier and when setting the altitude. The heating indicator lights when the humidifier is on.  CAUTION • • • Be careful to place the device where it cannot be bumped, and where no one will trip over the power cord. If you put the device on the floor, make sure the area is free from dust and clear of bedding, clothes or other objects that could block the air inlet. Make sure the area around the device is dry and clean. WARNING • • • Make sure the power cord and plug are in good condition and the equipment is not damaged. Only ResMed air tubing should be used with the device. A different type of air tubing may alter the pressure you actually receive and reduce the effectiveness of your treatment. Blocking the hose and/or air inlet of the device while in operation could lead to overheating of the device. 2 1 3 4 1. 2. 3. 4. Connect the power cord to the rear of the Tango. Plug the other end into a power outlet. Connect the air tubing to the air outlet of the Tango. Connect the assembled mask system to the free end of the air tubing.  English Setting up the Tango Using the Tango Start therapy when the Tango is in standby mode. A circulating dash symbol displays on the Press LCD. If ramp is set, the ramp time displays on the LCD. When ramping is finished, therapy continues at the prescribed pressure. If no ramp is set, therapy starts immediately at the prescribed pressure. Stop therapy Press Set ramp time 1. . Repeatedly press until the LCD displays the desired time. to accept changes and start therapy. (The Tango returns to standby mode Press after 5 seconds if no button is pressed.) The ramp settings are (in minutes): Off, 10, 20, 45. 2. Set altitude 1. Hold 2. Press down for 3 seconds until the current altitude setting is displayed. or to change the altitude setting. to accept changes and return to standby mode. (The Tango returns to Press standby mode after 10 seconds if no button is pressed.) The altitude settings are: 1 (0–2000’/0–609 m), 2 (2001–4000’/610–1219 m), 3 (4001–6000’/1220–1828 m), 4 (6001–8500’/1829–2591 m). 3. Cleaning and Maintenance WARNING Beware of electric shock. Do not immerse the flow generator or power cord in water. Always unplug the flow generator before cleaning and make sure it is dry before reconnecting. CAUTION • • Do not hang the air tubing in direct sunlight as it may harden over time and eventually crack. Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing or antibacterial soaps or scented oils to clean the air tubing or the Tango. These solutions may cause hardening and reduce the life of the product. Daily: Disconnect the air tubing and hang it in a clean, dry place until next use. Weekly: 1. Remove the air tubing and wash it in warm water and mild detergent. Rinse thoroughly, hang and allow to dry. 2. Before next use, reassemble the mask and headgear. 3. Reconnect the air tubing. Monthly: • Clean the exterior of the Tango with a damp cloth. • Check the air filter for holes and blockage by dirt. Replace it every six months, or more often in a dusty environment. For information on cleaning your mask or humidifier, refer to the manual supplied with your mask or humidifier.  English Changing the Air Filter Air inlet Air filter CAUTION Do not wash the air filter. 1. With the power cord and air tubing disconnected, turn the Tango upside down so the air inlet is facing you. The air filter is inside the air inlet. 2. Pull the old air filter out of the air inlet. 3. Put the new air filter into the air inlet in the direction shown by the arrow, with the blue side facing towards you. Troubleshooting Problem Possible Cause Solution Tango won’t start Power cord not connected properly. Check power cord. Faulty power outlet. Try another power outlet. The air tubing is not connected properly. Adjust the air tubing connection. The air tubing is blocked, pinched or punctured. Unblock or free the air tubing. Check the air tubing for punctures. Treatment pressure seems either low or too high Altitude setting incorrect. Adjust altitude setting. LCD displays an error message and treatment stops or can’t be started Problem with motor. Contact your supplier. No/low air flow  Technical Specifications Operating pressure range 4 to 20 cm H2O Dimensions (L x W x D) 8.5” x 7.4” x 4.5” (215 mm x 189 mm x 114 mm) Weight 2.4 lb (1.1 kg) Power supply Input range 110V, 60Hz, 20VA/10W (typical power consumption), 36VA/15W (maximum power consumption) Operating temperature +41°F to +104°F (+5°C to +40°C) Operating humidity 10–95% non-condensing Storage and transport temperature -4°F to +140°F (-20°C to +60°C) Storage and transport humidity 10–95% non-condensing Atmospheric pressure range 1060 hPa (sea level) to 680 hPa (8500’/ 2591 m) Electromagnetic Compatibility Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial and light industry environments Air Filter Two-layered, powder-bonded, polyester non-woven fiber Air Tubing Flexible plastic, 6’6” (2 m) Air Outlet The 0.87” (22 mm) conical air outlet complies with EN 1281-1 IEC 60601-1 Classification Class II (double insulation), Type CF The manufacturer reserves the right to change these specifications without notice. Symbols which appear on the Tango Attention. Consult accompanying documents; Type CF equipment; equipment; Drip proof; Canadian Standards Association. IPX1  Class II A caution explains special measures for safe and effective use of the device. • At low pressures, the flow through the exhalation ports of your mask may not clear all exhaled gas from the tubing. Some rebreathing may occur. • The airflow for breathing produced by this device can be as much as 11°F (6°C) higher than the temperature of the room. Be careful if the room temperature is warmer than 90°F (32°C).  Ports may be incorporated into the mask or in connectors that are near the mask.  English General Warnings and Cautions A warning alerts you to possible injury. • Read the entire guide before using the Tango. • The advice in this guide should not supersede instructions given by the prescribing physician. • The Tango should only be used with masks (and connectors*) recommended by ResMed, or by a physician or respiratory therapist. A mask should not be used unless the Tango is turned on and operating properly. The vent hole or holes associated with the mask should never be blocked. Explanation: The Tango is intended for use with special masks (or connectors) with vent holes allowing a continuous flow of air out of the mask. • When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask vent holes. If the device is not operating, insufficient fresh air is provided through the mask, and the exhaled air may be rebreathed. Rebreathing exhaled air for longer than several minutes can sometimes lead to suffocation. If there is a power failure or the device machine malfunctions, remove the mask. • The Tango can be set to deliver pressures up to 20 cm H2O. In the unlikely event of certain fault conditions, pressures up to 30 cm H2O are possible. • If oxygen is used with this device, the oxygen flow must be turned off when the device is not operating. If oxygen has been left on while the device is not operating, disconnect the oxygen and wait 30 minutes before turning on the flow generator again. Explanation: When the CPAP device is not in operation and the oxygen flow is left on, oxygen delivered into the air delivery tubing may accumulate within the CPAP machine enclosure and create a risk of fire. • Oxygen supports combustion. Do not use oxygen while you are smoking or in the presence of an open flame. • Always ensure air is flowing from the device before turning on the oxygen supply. • Always turn the oxygen supply off before stopping the airflow from the device. Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will vary, depending on where the oxygen is introduced, pressure settings, patient breathing pattern, mask selection and leak rate. • Explosion hazard—do not use in the vicinity of flammable anesthetics. • Do not use the Tango if there are obvious external defects or unexplained changes in performance. • Do not open the Tango case. There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an authorized service agent. Servicing The Tango should be inspected by an authorized ResMed service center 5 years from the date of manufacture. Before this, the device is intended to provide safe and reliable operation if it is operated and maintained according to the instructions provided by ResMed. Warranty details are provided with the device at the time of original supply. As with all electrical devices, if any irregularity becomes apparent, you should have the device inspected by an authorized ResMed service center. Limited Warranty ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable. Product Warranty Period ResMed humidifiers, ResControl™, ResLink™, ResTraxx™ 1 Year ResMed flow generators 2 Years Accessories, water chambers and mask systems (including mask frame, cushion, headgear and tubing). Excludes single-use devices. 90 Days Note: Some models are not available in all regions. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product; b) repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs; c) any damage or contamination due to cigarette, pipe, cigar or other smoke; d) any damage caused by water being spilled on or into a flow generator. Warranty is void on product sold, or resold, outside the region of original purchase. Warranty claims on defective product must be made by the initial consumer at the point of purchase. This warranty is in lieu of all other express or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you. ResMed shall not be responsible for any incidental or consequential damages claimed to have occurred as a result of the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from region to region. For further information on your warranty rights, contact your local ResMed dealer or ResMed office.  Guidance and manufacturer’s declaration—electromagnetic emissions The C-Series Tango system is intended for use in the electromagnetic environment specified below. The customer or the user of the C-Series Tango system should assure that the device is used in such an environment. Emissions test Compliance Electromagnetic environment—guidance RF emissions CISPR11 Group 1 The C-Series Tango system uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emissions CISPR 11 Class B Harmonic Emissions IEC 61000-3-2 Class A The C-Series Tango system is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage network that supplies buildings used for domestic purposes. Voltage Fluctuations/Flicker Emissions IEC 61000-3-3 Complies Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided in this document. Warnings: The C-Series Tango system should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the C-Series Tango system should be observed to verify normal operation in the configuration in which it will be used. The use of accessories (eg, humidifiers) other than those specified in this manual is not recommended. They may result in increased emissions or decreased immunity of the C-Series Tango system.  English Guidance and Manufacturer’s Declaration - Electromagnetic Emissions and Immunity Guidance and manufacturer’s declaration—electromagnetic immunity The C-Series Tango system is intended for use in the electromagnetic environment specified below. The customer or the user of the C-Series Tango system should assure that the device is used in such an environment. Immunity test IEC60601-1-2 test level Compliance level Electromagnetic environment— guidance Electrostatic discharge (ESD) IEC 61000-4-2 ±6 kV contact ±6 kV contact ±8 kV air ±8 kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Electrical fast transient/burst IEC 61000-4-4 ±2 kV for power supply lines ±2 kV ±1 kV for input/output lines Not Applicable Surge IEC 61000-4-5 ±1 kV differential mode ±1 kV differential mode ±2 kV common mode ±2 kV common mode <5% Ut (>95% dip in Ut) for 0.5 cycle <12V (>95% dip in 240V) for 0.5 cycle 40% Ut (60% dip in Ut) for 5 cycles 96V (60% dip in 240V) for 5 cycles 70% Ut (30% dip in Ut) for 25 cycles 168V (30% dip in 240V) for 25 cycles <5% Ut (>95%dip in Ut) for 5 sec <12V (>95%dip in 240V) for 5 sec C-Series Tango system be powered from an uninterruptible power source. 3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. Voltage dips, short interruptions and voltage variations on power supply input lines. IEC 61000-4-11 Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 Note: Ut is the AC mains voltage prior to application of the test level. 10 Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. If the user of the CSeries Tango system requires continued operation during power mains interruptions, it is recommended that the The C-Series Tango system is intended for use in the electromagnetic environment specified below. The customer or the user of the C-Series Tango system should assure that the device is used in such an environment. Immunity test IEC60601-1-2 test level Compliance level Electromagnetic environment—guidance Portable and mobile RF communications equipment should be used no closer to any part of the C-Series Tango system, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 3 Vrms 150 kHz to 80 MHz 3 Vrms d = 1.17 √P Radiated RF IEC 61000-4-3 10 V/m 80 MHz to 2.5 GHz 10 V/m d = 0.35 √P 80 MHz to 800 MHz d = 0.70 √P 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the C-Series Tango system is used exceeds the applicable RF compliance level above, the C-Series Tango system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the C-Series Tango system. a b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m. 11 English Guidance and manufacturer’s declaration—electromagnetic immunity Recommended separation distances between portable and mobile RF communications equipment and the C-Series Tango system The C-Series Tango system is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the C-Series Tango system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the C-Series Tango system as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter (m) Rated maximum output power of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz d = 1.17 √P d = 0.35 √P d = 0.7 √P 0.01 0.17 0.04 0.07 0.1 0.37 0.11 0.22 1 1.17 0.35 0.7 10 3.69 1.11 2.21 100 11.70 3.50 7.0 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. 12 CPAP Tango™ C-Series Guía del usuario Uso indicado PRECAUCIÓN La ley federal estipula que este aparato solamente puede ser vendido por un médico o por orden de éste. Contraindicación El tratamiento con presión positiva en las vías respiratorias puede estar contraindicado en algunos pacientes que presenten las siguientes afecciones previas: • enfermedad bullosa pulmonar grave • neumotórax • presión arterial patológicamente baja • deshidratación • fuga de líquido cefalorraquídeo, cirugía o traumatismo craneanos recientes. Efectos secundarios Los pacientes deben informar al médico tratante acerca de todo dolor torácico inusual, un dolor de cabeza severo o un aumento en su dificultad respiratoria. Una infección respiratoria alta aguda puede requerir la suspensión temporal del tratamiento. Los siguientes efectos secundarios pueden surgir durante el curso del tratamiento con el Tango: • sequedad de la nariz, boca o garganta • epistaxis • hinchazón por tragar aire • molestias sinusales o del oído • irritación ocular • erupciones cutáneas. 13 Español El sistema Tango CPAP es indicado para el tratamiento de la apnea obstructiva del sueño (AOS) en pacientes que pesen más de 30 kg (66 lb). Está diseñado para uso hospitalario y domiciliario. Piezas del Tango El Tango consta de: • Generador de aire Tango C-Series (aparece a continuación) • Cable de alimentación • Tubo de aire de 2 m (6’6”). Enchufe de cable de señal (sólo para uso con humidificador) Enchufe de cable de alimentación Salida de aire Entrada de aire (contiene filtro de aire) Mascarillas Los sistemas de mascarillas ResMed recomendados para ser utilizados con el Tango son: • Mascarilla facial Mirage™ Serie II • Mascarilla nasal Meridian™ • Mascarilla facial Ultra Mirage™ • Mascarilla nasal Mirage™ • Sistema de almohadillas nasales Mirage Swift™ • Mascarilla nasal Mirage Activa™ • Sistema de almohadillas nasales Mirage Swift II • Mascarilla nasal Mirage Vista™ • Mascarilla nasal Papillon™ • Mascarilla nasal Ultra Mirage™ II • Mascarilla nasal Silent Papillon™ • Mascarilla facial Mirage Liberty™ • Mascarilla facial Mirage Quattro™ Para información sobre el uso de las mascarillas, consulte el manual de su mascarilla. Humidificadores Para obtener información acerca del uso del humidificador térmico C-Series de Resmed o del ResMed Passover, consulte el manual del humidificador. 14 La interfaz del Tango Español Indicador de calefacción Indicador de alimentación Modo de espera Cuando se enciende el Tango, se enciende el indicador de alimentación y las horas de funcionamiento aparecen en el visor. Todos los cambios a las configuraciones de altitud y rampa se hacen en el modo de espera. Rampa Configurar un tiempo de rampa permite que el tratamiento comience con una presión baja y aumente hasta llegar a presión completa durante el tiempo de rampa (consulte “Configurar tiempo de rampa”). Altitud Los cambios de altitud inciden en la presión suministrada por el Tango. La función de altitud compensa los cambios de presión relativos a cambios de altitud (consulte “Configurar altitud”). Horas de funcionamiento Las horas de funcionamiento aparecen cuando el Tango está en modo de espera. Humidificador Los botones del humidificador se utilizan cuando se usa el humidificador térmico C-Series y al configurar la altitud. El indicador de calefacción se ilumina cuando está encendido el humidificador. Instalación del Tango PRECAUCIÓN • • • Tenga cuidado de no colocar el equipo en un lugar donde pueda ser golpeado o donde alguien pueda tropezar con el cable de alimentación. Si coloca el equipo en el piso, asegúrese de que el área esté libre de polvo y que no haya ropa, ropa de cama u otros objetos que pudieran obstruir la entrada de aire. Asegúrese de que el área alrededor del equipo esté limpia y seca. ADVERTENCIA • • • Asegúrese de que el cable de alimentación y el enchufe estén en buenas condiciones y que el equipo no esté dañado. Sólo debe usarse un tubo de aire ResMed con el equipo. Un tubo de aire diferente podría modificar la presión que de hecho recibe, lo que reduciría la eficacia del tratamiento. Bloquear la manguera y/o la entrada de aire del dispositivo mientras está en operación podría producir un sobrecalentamiento del dispositivo. 15 2 1 3 1. 2. 3. 4. 4 Conecte el cable de alimentación a la parte trasera del Tango. Enchufe el otro extremo a una toma de corriente. Conecte el tubo de aire a la salida de aire del Tango. Conecte el sistema de mascarilla armado al extremo libre del tubo de aire. Uso del Tango Inicio del tratamiento cuando el Tango esté en modo de espera. Aparece un símbolo de guión Presione circulante en el visor. Si está configurada la rampa, aparece en el visor el tiempo de rampa. Cuando concluye la rampa, el tratamiento sigue con la presión prescrita. Si no se ha configurado la rampa, el tratamiento comienza de inmediato con la presión prescrita. Detención del tratamiento Presione Configurar el tiempo de rampa 1. Configurar altitud 1. presionado durante 3 segundos hasta que aparezca la configuración Mantenga de altitud actual. 2. Para cambiar la configuración de altitud, presione . Presione varias veces hasta que en el visor aparezca el tiempo que desee. para aceptar los cambios y comenzar el tratamiento. (El Tango vuelve al Presione modo de espera después de 5 segundos si no se presiona ningún botón.) Las configuraciones de rampa son (en minutos): Apag, 10, 20, 45. 2. o . para aceptar los cambios y volver al modo de espera. (El Tango vuelve al Presione modo de espera después de 10 segundos si no se presiona ningún botón.) Las configuraciones de altitud son: 1 (0–609 m/0–2000 pies) 2 (610–1219 m/2001–4000 pies) 3 (1220–1828 m/4001–6000 pies) 4 (1829–2591 m/6001–8500 pies). 3. 16 Limpieza y mantenimiento ADVERTENCIA Peligro de electrocución. No sumerja en agua el generador de aire ni tampoco el cable de alimentación. Desconecte siempre el generador de aire antes de limpiarlo y asegúrese de que esté seco antes de volver a conectarlo. • No cuelgue el tubo de aire a la luz solar directa, ya que puede endurecerse y agrietarse con el tiempo. No utilice blanqueadores, cloro, alcohol, soluciones aromáticas, jabones humectantes ni antibacterianos ni aceites aromatizados para limpiar el tubo de aire del Tango. Estas soluciones podrían endurecer el producto y reducir la vida útil del mismo. Diariamente: Desconecte el tubo de aire y cuélguelo en un lugar limpio y seco hasta el próximo uso. Semanalmente: 1. Saque el tubo de aire y lávelo con agua tibia y detergente suave. Enjuáguelo bien, cuélguelo y déjelo secar. 2. Antes de usarlo otra vez, vuelva a armar la mascarilla y el arnés para la cabeza. 3. Vuelva a conectar el tubo de aire. Mensualmente • Limpie el exterior del Tango con un paño húmedo. • Verifique que el filtro de aire no tenga orificios ni esté obstruido por suciedad. Reemplácelo cada seis meses, o con mayor frecuencia si se usa en un entorno con mucho polvo. Para más información sobre la limpieza de la mascarilla o del humidificador, consulte el manual suministrado con la mascarilla o con el humidificador. 17 Español PRECAUCIÓN • Cambio del filtro de aire Entrada de aire Filtro de aire PRECAUCIÓN No lave el filtro de aire. 1. Con el cable de alimentación y el tubo de aire desconectados, voltee el Tango de manera que la entrada de aire se oriente hacia usted. El filtro de aire está dentro de la entrada de aire. 2. Saque el filtro de aire usado de la entrada de aire. 3. Coloque el filtro de aire nuevo en la entrada de aire en la dirección que indica la flecha, con el lado azul orientado hacia usted. Solución de problemas Problema Causa posible Solución El Tango no se pone en marcha El cable de alimentación no está debidamente conectado. Revise el cable de alimentación. La toma de corriente está defectuosa. Pruebe otra toma de corriente. El tubo de aire no está debidamente conectado. Ajuste la conexión del tubo de aire. El tubo de aire está obstruido, torcido o roto. Desobstruya o despeje el tubo de aire. Verifique que el tubo de aire no esté perforado. La presión de tratamiento parece ser demasiado baja o demasiado alta Configuración de altitud incorrecta. Ajuste la configuración de altitud. El visor muestra un mensaje de error y el tratamiento se detiene o no puede ponerse en marcha Problema con el motor. Póngase en contacto con su proveedor. No hay flujo de aire o éste es muy bajo 18 Especificaciones técnicas Rango de presión de funcionamiento: 4 a 20 cm H2O Dimensiones (largo x ancho x altura): 215 mm x 189 mm x 114 mm (8,5 pulgadas x 7,4 pulgadas x 4,5 pulgadas) 1,1 kg (2,4 lb) Rango de entrada 110 V, 60 Hz, 20 VA/10 W (consumo de energía típico), 36 VA/15 W (máximo consumo de energía) Temperatura de funcionamiento +5°C a +40°C (+41°F a +104°F) Humedad de funcionamiento 10–95% sin condensación Temperatura de almacenamiento y transporte: -20°C a +60°C (-4°F a +140°F) Humedad de almacenamiento y transporte: 10–95% sin condensación Rango de presión atmosférica 1060 hPa (a nivel del mar) a 680 hPa (2591 m/8500’) Compatibilidad electromagnética El producto cumple con todos los requisitos pertinentes en cuanto a compatibilidad electromagnética (EMC por sus siglas en inglés) de acuerdo con la CEI 60601-1-2, para entornos residenciales, comerciales y de industria ligera. Filtro de aire Dos capas de fibra de poliéster no entrelazada, con unión de polvo Español Peso Alimentación Tubo de aire Plástico flexible, 2 m (6 pies y 6 pulgadas) Salida de aire La salida de aire cónica de 22 mm (0,87 pulgadas) cumple con la norma EN 1281-1 Clasificación CEI 60601-1 Clase II (doble aislamiento), Tipo CF El fabricante se reserva el derecho de cambiar estas especificaciones sin previo aviso. Símbolos que aparecen en el Tango IPX1 Atención. Consulte los documentos adjuntos; Equipo tipo CF; A prueba de goteo; Asociación canadiense de estándares. 19 Equipo de clase II; Advertencias y precauciones generales Una advertencia alerta sobre la posibilidad de que se produzcan lesiones. • Lea por completo la guía antes de usar el Tango. • Las indicaciones de esta guía no deben suplantar las instrucciones entregadas por el médico tratante. • El Tango sólo debe usarse con mascarillas (y conectores*) recomendadas por ResMed, o por un médico o terapeuta respiratorio. No debe usarse una mascarilla a menos que el Tango esté encendido y funcionando correctamente. Nunca se debe obstruir el o los orificios de ventilación asociados con la mascarilla. Explicación: El Tango está diseñado para ser usado con mascarillas (o conectores) especiales con orificios de ventilación que permiten un flujo continuo de aire desde la mascarilla. • Mientras el equipo está encendido y funcionando correctamente, el aire fresco del dispositivo desplaza el aire espirado hacia fuera de la mascarilla a través de sus orificios de ventilación. Si el dispositivo no está funcionando, no se proporciona suficiente aire fresco por la mascarilla, y es posible que se vuelva a respirar aire exhalado. Volver a respirar aire exhalado durante varios minutos puede en ocasiones conducir a la asfixia. Si se produce una falla eléctrica o si falla el dispositivo, retire la mascarilla. • El Tango puede configurarse para suministrar presiones de hasta 20 cm H2O. En el caso poco probable de ciertas situaciones de fallas, es posible que se suministren presiones de hasta 30 cm H2O. • Si se utiliza oxígeno con este equipo, el flujo del mismo deberá estar cerrado mientras el equipo no esté en funcionamiento. Si se deja activado el oxígeno mientras el dispositivo no está funcionando, desconecte el oxígeno y espere 30 minutos antes de volver a encender el generador de aire. Explicación: mientras el equipo CPAP no esté funcionando, si se deja abierto el flujo de oxígeno, el oxígeno suministrado al tubo de respiración podrá acumularse dentro del dispositivo CPAP, creando así un riesgo de incendio. • El oxígeno favorece la combustión. No utilice oxígeno mientras esté fumando ni en presencia de llamas expuestas. • Asegúrese siempre de que esté fluyendo aire desde el dispositivo antes de encender el suministro de oxígeno. • Siempre apague el suministro de oxígeno antes de detener el flujo de aire del equipo. Nota: A un determinado caudal de oxígeno complementario, la concentración de oxígeno inhalado variará según dónde se introduzca el oxígeno, los parámetros de presión, el ritmo respiratorio del paciente, la selección de la mascarilla y el índice de fuga. • Peligro de explosión: no utilizar cerca de anestésicos inflamables. • No utilice el Tango si presenta defectos externos obvios o si ocurren cambios inexplicados en su funcionamiento. • No abra la caja del Tango. No hay piezas dentro que el usuario pueda reparar. Las reparaciones y el mantenimiento interno deberán ser efectuados únicamente por un agente de servicio autorizado.  Pueden integrarse puertos en la mascarilla o en los conectores próximos a la mascarilla. 20 Servicio El Tango debe ser inspeccionado por un centro de servicio autorizado de ResMed 5 años después de la fecha de fabricación. Antes de esto, el dispositivo está diseñado para proporcionar un funcionamiento seguro y confiable si se opera y mantiene de acuerdo con las instrucciones provistas por ResMed. Los detalles acerca de la garantía acompañan el dispositivo al momento de la entrega original de éste. Al igual que con todos los dispositivos eléctricos, si se hace evidente la presencia de una irregularidad, debe hacer revisar el dispositivo por un centro de servicio autorizado de ResMed. Garantía limitada ResMed garantiza que su producto ResMed estará libre de todo defecto de material y mano de obra durante el plazo estipulado a continuación a partir de la fecha de adquisición por parte del consumidor inicial. Esta garantía no es transferible. Producto Plazo de la garantía Humidificadores ResMed, ResControl™, ResLink™, ResTraxx™ 1 año Generadores de aire ResMed 2 años Accesorios, cámaras de agua y sistemas de mascarillas (incluido el armazón de mascarilla, la almohadilla, el arnés para la cabeza y el tubo). Excluye los dispositivos desechables. 90 días Nota: Algunos modelos no se encuentran disponibles en todas las regiones. Si el producto falla bajo condiciones de utilización normales, ResMed reparará o reemplazará, a opción suya, el producto defectuoso o cualquiera de sus componentes. Esta garantía limitada no cubre: a) cualquier daño causado como resultado de una utilización indebida, abuso, modificación o alteración del producto; b) reparaciones llevadas a cabo por cualquier organización de servicio que no haya sido expresamente autorizada por ResMed para efectuar dichas reparaciones; c) cualquier daño o contaminación debido a humo de cigarrillo, pipa, cigarro u otras fuentes de humo; d) cualquier daño causado por el derrame de agua sobre el generador de aire o hacia su interior. La garantía queda anulada si el producto se vende o revende fuera de su región de compra original. Las reclamaciones de garantía con respecto a productos defectuosos deben ser realizadas por el consumidor original en el punto de compra. 21 Español Una precaución explica medidas especiales para el uso seguro y eficaz del dispositivo. • A bajas presiones, el flujo a través de los puertos de exhalación de la mascarilla puede no eliminar todos los gases exhalados del tubo. Puede darse cierto grado de reinspiración. • El flujo de aire generado por este dispositivo para la respiración puede llegar a estar hasta 6°C (11°F) por encima de la temperatura ambiente. Tenga cuidado si la temperatura ambiente es de más de 32°C (90°F). Esta garantía revoca cualquier otra garantía expresa o implícita, incluso cualquier garantía implícita de comerciabilidad o idoneidad para un propósito en particular. Algunas regiones o estados no permiten limitaciones respecto a la duración de una garantía implícita, por lo que es posible que lo que precede no sea aplicable en su caso. ResMed no será responsable de ningún daño incidental ni emergente que se reclame por haber ocurrido como resultado de la venta, instalación o utilización de cualquier producto de ResMed. Algunas regiones o estados no permiten la exclusión ni limitación de daños incidentales o emergentes, por lo que es posible que lo que precede no sea aplicable en su caso. Esta garantía le otorga derechos legales específicos, y es posible que usted tenga otros derechos que pueden variar de una región a otra. Para más información sobre los derechos que le otorga esta garantía, póngase en contacto con su distribuidor local de ResMed o con una oficina de ResMed. Guía y declaración del fabricante – Emisiones e inmunidad electromagnéticas Guía y declaración del fabricante – emisiones electromagnéticas El sistema Tango C-Series está diseñado para ser utilizado en el entorno electromagnético especificado a continuación. El cliente o usuario del sistema Tango C-Series deberá asegurarse de que sea utilizado en dicho entorno. Prueba de emisiones Cumplimiento Entorno electromagnético – guía Emisiones de RF CISPR11 Grupo 1 El sistema Tango C-Series usa energía de RF sólo para su funcionamiento interno. Por lo tanto, sus emisiones de RF son muy bajas y es improbable que provoquen interferencias en equipos electrónicos cercanos. Emisiones de RF CISPR11 Clase B Emisiones armónicas CEI 61000-3-2 Clase A El sistema Tango C-Series es apropiado para ser usado en todos los ámbitos, incluidos ámbitos domésticos y aquéllos conectados directamente a la red pública de bajo voltaje que abastece a los edificios destinados a vivienda. Fluctuaciones de voltaje / Emisiones de parpadeo (flicker) CEI 61000-3-3 Cumple Un equipo médico eléctrico requiere precauciones especiales en lo concerniente a la compatibilidad electromagnética, y debe ser instalado y puesto en funcionamiento según la información relativa a dicha compatibilidad que se proporciona en este documento. Advertencias: El sistema Tango C-Series no debe utilizarse al lado de otro equipo ni montado sobre o debajo de él. Si dicha proximidad o el tenerlo montado sobre o debajo de otro equipo fueran inevitables, el sistema Tango C-Series debería ser observado para verificar su normal funcionamiento en la configuración en la que será utilizado. No se recomienda el uso de accesorios (por ej.: humidificadores) distintos a los que se especifican en este manual. Dichos humidificadores podrían provocar un aumento de las emisiones o una disminución de la inmunidad del sistema Tango C-Series. 22 Guía y declaración del fabricante – inmunidad electromagnética Prueba de inmunidad Nivel de prueba CEI 60601-1-2 Nivel de cumplimiento Entorno electromagnético – guía Descarga electrostática CEI 61000-4-2 ± 6 kV por contacto ± 6 kV por contacto ± 8 kV al aire ± 8 kV al aire Los pisos deben ser de madera, cemento o baldosas de cerámica. Si los pisos están revestidos de material sintético, la humedad relativa debería ser al menos 30%. Transitorios eléctricos rápidos/en ráfagas CEI 61000-4-4 ± 2 kV para líneas de suministro de energía ± 2 kV ± 1 kV para líneas de entrada y de salida No corresponde Subida rápida de tensión CEI 61000-4-5 ± 1 kV en modo diferencial ± 1 kV en modo diferencial ± 2 kV en modo común ± 2 kV en modo común <5% Ut (>95% de caída con respecto al Ut) durante 0,5 ciclos <12V (>95% de caída en 240 V) durante 0,5 ciclos 40% del Ut (60% de caída con respecto al Ut) durante 5 ciclos 96 V (60% de caída en 240 V) durante 5 ciclos 70% del Ut (30% de caída con respecto al Ut) durante 25 ciclos 168 V (30% de caída en 240 V) durante 25 ciclos <5% Ut (>95% de caída con respecto al Ut) durante 5 seg <12 V (>95% de caída en 240 V) durante 5 seg 3 A/m 3 A/m Caídas de voltaje, variaciones de voltaje e interrupciones cortas en las líneas de entrada del suministro de energía CEI 61000-4-11 Campo magnético de la frecuencia industrial (50/60 Hz) CEI 61000-4-8 Nota: Ut es la tensión de red de CA antes de la aplicación del nivel de prueba. 23 La calidad de la red de energía eléctrica deberá ser la de un típico entorno comercial u hospitalario. La calidad de la red de energía eléctrica deberá ser la de un típico entorno comercial u hospitalario. La calidad de la red de energía eléctrica deberá ser la de un típico entorno comercial u hospitalario. Si el usuario del sistema Tango C-Series necesita un funcionamiento continuo a pesar de cortes en el suministro de energía, se recomienda que el sistema Tango C-Series sea alimentado por una fuente de energía continua. Los campos magnéticos de la frecuencia industrial deben tener los niveles propios de un local típico de un entorno comercial u hospitalario típico. Español El sistema Tango C-Series está diseñado para ser utilizado en el entorno electromagnético especificado a continuación. El cliente o usuario del sistema Tango C-Series deberá asegurarse de que sea utilizado en dicho entorno. Guía y declaración del fabricante – inmunidad electromagnética El sistema Tango C-Series está diseñado para ser utilizado en el entorno electromagnético especificado a continuación. El cliente o usuario del sistema Tango C-Series deberá asegurarse de que sea utilizado en dicho entorno. Prueba de inmunidad CEI60601-1-2 Nivel de prueba Nivel de cumplimiento Entorno electromagnético – guía Los equipos de comunicaciones por RF portátiles y móviles no deben ser usados cerca de parte alguna del sistema Tango C-Series (incluidos los cables) a una distancia menor que la recomendada, calculada a partir de la ecuación aplicable a la frecuencia del transmisor. Distancia recomendada RF conducida CEI 61000-4-6 3 Vrms de 150 kHz a 80 MHz 3 Vrms d = 1,17 √P RF irradiada CEI 61000-4-3 10 V/m de 80 MHz a 2,5 GHz 10 V/m d = 0,35 √P 80 MHz a 800 MHz d = 0,70 √P 800 MHz a 2,5 GHz Donde P es la potencia nominal máxima de salida del transmisor en vatios (W) según el fabricante del transmisor y d es la distancia recomendada en metros (m). Las intensidades de campo de transmisores de RF fijos, determinadas según un control electromagnético del lugar, a deben ser menores que el nivel de cumplimiento en cada rango de frecuencias.b Puede haber interferencias cerca de equipos marcados con el siguiente símbolo: Nota 1: A 80 MHz y 800 MHz se aplica el rango de frecuencias más alto. Nota 2: Es posible que estas guías no se apliquen en todas las situaciones. La propagación electromagnética es afectada por la absorción y el reflejo en estructuras, objetos y personas. Las intensidades de campo de los transmisores fijos, tales como estaciones base para teléfonos de radio (celulares/inalámbricos) y radios móviles terrestres, radio amateur, transmisores de radio AM y FM y de televisión, no se pueden predecir con exactitud en forma teórica. Para evaluar el entorno electromagnético provocado por transmisores de RF fijos, se debería considerar la realización de un control electromagnético del lugar. Si la intensidad de campo medida en el lugar donde habrá de usarse el sistema Tango C-Series excediera el nivel de cumplimiento de RF correspondiente indicado antes, el sistema Tango C-Series debería ser vigilado en cuanto a su normal funcionamiento. Si ocurriera un funcionamiento anormal, pueden ser necesarias medidas adicionales, como la reorientación o reubicación del sistema Tango C-Series. a b En el rango de frecuencias de 150 kHz a 80 MHz, las intensidades de campo deberían ser menores de 10 V/m. 24 Distancias recomendadas entre equipos de comunicación por RF portátiles y móviles y el sistema Tango C-Series El sistema Tango C-Series está diseñado para ser usado en un entorno en el que las alteraciones por RF irradiada sean controladas. El cliente o usuario del sistema Tango C-Series puede ayudar a prevenir la interferencia electromagnética guardando una distancia mínima entre los equipos de comunicaciones por RF portátiles y móviles (transmisores) y el sistema Tango C-Series, tal como se recomienda a continuación según la potencia máxima de salida del equipo de comunicaciones. 150 kHz a 80 MHz 80 MHz a 800 MHz 800 MHz a 2,5 GHz d = 1,17 √P d = 0,35 √P d = 0,7 √P 0,01 0,17 0,04 0,07 0,1 0,37 0,11 0,22 1 1,17 0,35 0,7 10 3,69 1,11 2,21 100 11,70 3,50 7,0 Para transmisores con una potencia nominal máxima de salida que no figure en la tabla precedente, la distancia recomendada d en metros (m) se puede determinar por medio de la ecuación que se aplica a la frecuencia del transmisor, donde P es la potencia nominal máxima de salida del transmisor en vatios (W), según el fabricante del transmisor. Nota 1: A 80 MHz y 800 MHz, se aplica la distancia para el rango de frecuencias más alto. Nota 2: Es posible que estas guías no se apliquen en todas las situaciones. La propagación electromagnética es afectada por la absorción y el reflejo en estructuras, objetos y personas. 25 Español Distancia según la frecuencia del transmisor (m) Potencia nominal máxima de salida del transmisor W 26 C-Series Tango™ Guide utilisateur Usage prévu L’appareil de PPC Tango est prévu pour le traitement du syndrome d’apnée obstructive du sommeil (SAOS) chez les patients de plus de 66 lb (30 kg). Il est prévu pour une utilisation à domicile et en milieu médical. ATTENTION Selon la loi fédérale américaine, ce dispositif ne peut être vendu aux U.S.A. que par un médecin ou sur prescription médicale. Le traitement par PPC est contre-indiqué chez certains patients présentant l’une des pathologies préexistantes suivantes : • affection pulmonaire bulleuse grave • pneumothorax • hypotension • déshydratation • fuite de liquide céphalo-rachidien, récent traumatisme crânien ou intervention chirurgicale crânienne. Effets secondaires Les patients doivent avertir leur médecin traitant en cas de douleurs thoraciques inhabituelles, de maux de tête sévères ou d’une dyspnée accrue. Une infection aiguë des voies aériennes supérieures peut nécessiter l’arrêt temporaire du traitement. Les effets secondaires suivants peuvent apparaître au cours du traitement sous Tango : • sécheresse nasale, buccale ou de la gorge • saignements de nez • ballonnements • gêne au niveau de l’oreille ou des sinus • irritation des yeux • érythèmes cutanés. 27 Français Contre-indications Composants du Tango Votre Tango se compose des éléments suivants : • Appareil C-Series Tango (représenté ci-dessous) • Câble d’alimentation • Circuit respiratoire 6’6” (2 m). Prise du câble d’interface (avec un humidificateur uniquement) Prise du câble d’alimentation Sortie d’air Entrée d’air (contient le filtre à air) Masques Les masques ResMed suivants sont recommandés pour utilisation avec le Tango : • Masque facial Mirage™ Série 2 • Masque nasal Meridian™ • Masque facial Ultra Mirage™ • Masque nasal Mirage™ • Masque narinaire Mirage Swift™ • Masque nasal Mirage Activa™ • Masque narinaire Mirage Swift II • Masque nasal Mirage Vista™ • Masque nasal Papillon™ • Masque nasal Ultra Mirage™ II • Masque nasal Silent Papillon™ • Masque facial Mirage Liberty™ • Masque facial Mirage Quattro™ Pour les informations concernant l’utilisation des masques, veuillez vous référer au manuel de votre masque. Humidificateurs Pour les informations concernant l’utilisation de l’humidificateur chauffant C-Series ResMed ou Passover ResMed, veuillez vous référer au manuel de votre humidificateur. 28 Interface du Tango Témoin de chauffe Mode d’attente Lorsque le Tango est sous tension, le témoin d’alimentation est allumé et les heures de fonctionnement sont affichées sur l’écran LCD. Toutes les modifications des réglages d’altitude et de rampe doivent être effectuées en mode d’attente. Rampe Le réglage d’une durée de rampe permet de démarrer le traitement à une pression basse, qui augmente progressivement jusqu’à la pression de traitement sur la durée de rampe (voir « Réglage de la durée de rampe »). Altitude Le changement d’altitude a une incidence sur la pression délivrée par le Tango. La fonction d’altitude permet de compenser les changements de pression liés aux changements d’altitude (voir « Réglage de l’altitude »). Heures de fonctionnement Les heures de fonctionnement sont affichées lorsque le Tango est en mode d’attente. Humidificateur Les touches de commande de l’humidificateur sont utilisées lorsque l’humidificateur chauffant C-Series est utilisé et lors du réglage de l’altitude. Le témoin de chauffe est allumé lorsque l’humidificateur est en marche. 29 Français Témoin d’alimentation Installation du Tango ATTENTION • • • Veiller à placer l’appareil de telle manière que personne ne puisse le faire tomber ou se prendre les pieds dans le câble d’alimentation. Si l’appareil est posé sur le sol, veiller à ce qu’il le soit dans un endroit exempt de poussière, de literie, de vêtements ou de tout autre objet susceptible d’obstruer l’entrée d’air. Vérifier que la zone autour de l’appareil est propre et sèche. AVERTISSEMENT • • • Vérifier que le câble d’alimentation et la fiche sont en bon état et que l’équipement n’est pas endommagé. Seul le circuit respiratoire ResMed peut être utilisé avec l’appareil. Les circuits respiratoires d’un autre type risquent de modifier la pression reçue et de réduire l’efficacité du traitement. L’obstruction du circuit ou de l’entrée d’air pendant le fonctionnement risque de causer une surchauffe de l’appareil. 2 1 3 1. 2. 3. 4. 4 Branchez le câble d’alimentation à l’arrière du Tango. Branchez l’autre extrémité du câble d’alimentation à une prise de courant. Raccordez le circuit respiratoire à la sortie d’air du Tango. Raccordez le masque assemblé à l’extrémité libre du circuit respiratoire. 30 Utilisation du Tango Démarrage du traitement lorsque le Tango est en mode d’attente. Un cercle en pointillés Appuyez sur la touche est affiché sur l’écran LCD. Si la rampe est réglée, sa durée est affichée sur l’écran LCD. Une fois qu’elle s’est écoulée, le traitement se poursuit à la pression prescrite. Si aucune rampe n’est réglée, le traitement démarre immédiatement à la pression prescrite. Arrêt du traitement Appuyez sur la touche Réglage de la durée de rampe 1. Appuyez plusieurs fois sur la touche l’écran LCD. jusqu’à ce que le réglage voulu s’affiche sur pour accepter les modifications de réglage et démarrer Appuyez sur la touche le traitement. (Le Tango retourne en mode d’attente au bout de 5 secondes si vous n’appuyez sur aucune autre touche.) Les options de réglage de rampe sont les suivantes (en minutes) : Off, 10, 20, 45. 2. 1. Maintenez la touche enfoncée pendant 3 secondes jusqu’à ce que le réglage d’altitude actuel s’affiche. 2. Appuyez sur la touche ou pour modifier le réglage d’altitude. pour accepter les modifications de réglage et retourner en Appuyez sur la touche mode d’attente. (Le Tango retourne en mode d’attente au bout de 10 secondes si vous n’appuyez sur aucune autre touche.) Les options de réglage pour l’altitude sont les suivantes : 1 (0 – 2000’ / 0 – 609 m), 2 (2001 – 4000’ / 610 – 1219 m), 3 (4001 – 6000’ / 1220 – 1828 m), 4 (6001 – 8500’ / 1829 – 2591 m). 3. Nettoyage et entretien AVERTISSEMENT Risque de chocs électriques. Ne pas immerger l’appareil ni le câble d’alimentation dans l’eau. Systématiquement débrancher l’appareil avant de le nettoyer et vérifier qu’il est bien sec avant de le brancher à nouveau. ATTENTION • • Ne pas suspendre le circuit respiratoire à la lumière directe du soleil afin d’éviter qu’il ne durcisse ou ne se fissure avec le temps. Ne pas utiliser d’eau de Javel, de chlore, d’alcool ou de solutions à base d’aromates, ni de savons hydratants ou antibactériens ou d’huiles parfumées pour le lavage du circuit respiratoire ou du Tango. Ces solutions peuvent causer le durcissement du produit et réduire sa durée de vie. Quotidien : détachez le circuit respiratoire et suspendez-le dans un endroit propre et sec jusqu’à la prochaine utilisation. Hebdomadaire : 1. Détachez le circuit respiratoire et nettoyez-le à l’eau tiède avec un détergent doux. Rincez soigneusement et suspendez-le pour le faire sécher. 2. Remontez le masque et le harnais avant de les réutiliser. 3. Raccordez le circuit respiratoire à nouveau. 31 Français Réglage de l’altitude . Mensuel : • Nettoyez l’extérieur du Tango à l’aide d’un chiffon humide. • Vérifiez que le filtre à air n’est ni percé ni bloqué par de la saleté. Remplacez-le tous les six mois ou plus souvent si l’appareil est utilisé dans un endroit poussiéreux. Pour les informations concernant le nettoyage du masque ou de l’humidificateur, veuillez vous référer au manuel fourni avec le masque ou l’humidificateur. Remplacement du filtre à air Entrée d’air Filtre à air ATTENTION Ne pas laver le filtre à air. 1. Avec le câble d’alimentation débranché et le circuit respiratoire détaché, mettez le Tango à l’envers afin que l’entrée d’air se trouve face à vous. Le filtre à air se trouve dans l’entrée d’air. 2. Tirez sur le filtre à air usagé pour le retirer de l’entrée d’air. 3. Mettez un nouveau filtre à air en place dans l’entrée d’air dans le sens indiqué par la flèche, avec la face bleue face à vous. Stratégie de dépannage Problème Cause possible Solution Le Tango ne démarre pas. Le câble d’alimentation n’est pas correctement branché. Vérifiez le câble d’alimentation. La prise de courant est défectueuse. Essayez une autre prise de courant. Le circuit respiratoire n’est pas correctement raccordé. Ajustez le raccordement du circuit respiratoire. Le circuit respiratoire est obstrué, pincé ou percé. Débouchez ou désentortillez le circuit respiratoire. Vérifiez que le circuit respiratoire n’est pas percé. La pression de traitement semble basse ou trop élevée. Réglage de l’altitude incorrect. Ajustez le réglage d’altitude. Un message d’erreur est affiché sur l’écran LCD et le traitement s’arrête ou ne démarre pas du tout. Problème de moteur. Contactez votre prestataire de santé. Absence de débit d’air/ Débit d’air faible. 32 Pression de fonctionnement 4 à 20 cm H2O Dimensions (L x l x P) 8,5” x 7,4” x 4,5” (215 mm x 189 mm x 114 mm) Poids 2,4 lb (1,1 kg) Alimentation électrique Courant 110 V, 60 Hz, 20 VA / 10 W (consommation typique de courant), 36 VA / 15 W (consommation maximale de courant) Température de fonctionnement +41°F à +104°F (+5°C à +40°C) Humidité de fonctionnement 10 à 95 % sans condensation Température de stockage et de transport -4°F à +140°F (-20°C à +60°C) Humidité de stockage et de transport 10 à 95 % sans condensation Fourchette de pressions atmosphériques 1060 hPa (niveau de la mer) à 680 hPa (8500’ / 2591 m) Compatibilité électromagnétique Le produit est conforme à toutes les exigences de compatibilité électromagnétique (CEM) applicables conformément à CEI 60601-1-2 pour les environnements résidentiels, commerciaux et l’industrie légère. Filtre à air Deux couches de mousse à cellules ouvertes de polyester, imprégnées de fibres polyester Circuit respiratoire Plastique flexible, longueur 6’6” (2 m) Sortie d’air Le raccord conique 0,87” (22 mm) de sortie d’air est conforme à la norme EN 1281-1 Classification CEI 60601-1 Classe II (double isolation), Type CF Le fabricant se réserve le droit de modifier ces caractéristiques sans notification préalable. Symboles apparaissant sur le Tango Attention. Veuillez consulter la documentation jointe ; Équipement de classe II ; Protection anti-gouttes ; Normalisation. IPX1 33 Équipement de type CF ; Association Canadienne de Français Caractéristiques Avertissements et précautions d’ordre général Un avertissement vous met en garde contre le risque de blessure. • Lire le manuel entièrement avant d’utiliser le Tango. • Les conseils donnés dans ce manuel ne remplacent pas les instructions du médecin traitant. • Le Tango doit être utilisé uniquement avec les masques (et raccords*) recommandés par ResMed, un médecin ou un kinésithérapeute respiratoire. Le masque ne peut être porté que si le Tango est sous tension et s’il fonctionne correctement. Les orifices de ventilation du masque ne doivent jamais être obstrués. Explication : le Tango est conçu pour être utilisé avec des masques (ou raccords) spéciaux possédant des orifices de ventilation qui permettent l’écoulement continu de l’air hors du masque. • Lorsque l’appareil est sous tension et qu’il fonctionne correctement, de l’air propre provenant de l’appareil expulse l’air expiré via les orifices de ventilation du masque. Si l’appareil ne fonctionne pas, le débit d’air propre délivré dans le masque est insuffisant et une réinhalation de l’air expiré peut se produire. La réinhalation de l’air expiré pendant une durée supérieure à quelques minutes peut parfois entraîner la suffocation. Le masque doit être retiré en cas de coupure de courant ou de dysfonctionnement de l’appareil. • Le Tango peut être réglé pour délivrer des pressions allant jusqu’à 20 cm H2O. Dans le cas, peu probable, d’une panne, les pressions peuvent s’élever jusqu’à 30 cm H2O. • Si de l’oxygène est utilisé avec cet appareil, l’alimentation en oxygène doit être fermée lorsque l’appareil n’est pas en mode de fonctionnement. Si l’alimentation en oxygène a été laissée ouverte avec l’appareil arrêté, déconnecter l’alimentation en oxygène, puis attendre 30 minutes avant de remettre l’appareil sous tension. Explication : lorsque l’appareil de PPC n’est pas en marche et que l’alimentation en oxygène n’est pas coupée, l’oxygène délivré dans le circuit respiratoire peut s’accumuler dans l’appareil et créer un risque d’incendie. • L’oxygène est inflammable. Ne pas fumer ni approcher de flamme nue de l’appareil lors de l’utilisation d’oxygène. • Systématiquement vérifier qu’un débit d’air est généré par l’appareil avant d’ouvrir l’alimentation en oxygène. • Systématiquement couper l’alimentation en oxygène avant d’arrêter le débit d’air de l’appareil. Remarque : lorsque de l’oxygène additionnel est délivré à un débit fixe, la concentration de l’oxygène inhalé varie en fonction du point d’entrée de l’oxygène, des réglages de pression, de la respiration du patient, du masque sélectionné ainsi que du niveau de fuite. • Danger d’explosion – Ne pas utiliser l’appareil à proximité d’anesthésiques inflammables. • Ne pas utiliser le Tango en cas de défauts externes évidents ou si sa performance devient aléatoire. • Ne pas ouvrir le boîtier du Tango. Il ne contient aucune pièce pouvant être entretenue par l’utilisateur. Les réparations et l’entretien interne doivent être effectués par un technicien agréé.  Des ports peuvent être intégrés au masque ou à des raccords proches du masque. 34 Une précaution (Attention) explique des mesures particulières permettant d’assurer une utilisation efficace et sûre de l’appareil. • À basses pressions, le débit aux orifices de ventilation de votre masque peut s’avérer insuffisant pour expulser tous les gaz expirés hors du circuit respiratoire. Une réinhalation peut se produire. • La température du débit d’air produit par cet appareil peut être supérieure de 11°F (6°C) à la température ambiante. Des précautions devront être prises lorsque la température ambiante est supérieure à 90°F (32°C). Le Tango doit être inspecté par un service technique agréé ResMed 5 ans après la date de fabrication. Avant cette date, l’appareil devrait assurer un fonctionnement fiable et sûr s’il est utilisé et entretenu conformément aux instructions fournies par ResMed. Les informations de garantie sont fournies avec l’appareil lors de sa livraison initiale. Comme avec tous les appareils électriques, si vous constatez un défaut quelconque, vous devez faire vérifier l’appareil par un service technique agréé ResMed. Garantie limitée ResMed garantit votre produit ResMed contre tout défaut de matériaux et de main-d’œuvre pour la période spécifiée ci-après à compter de la date d’achat par le client d’origine. Cette garantie n’est pas cessible. Produit Durée de la garantie Humidificateurs ResMed, ResControl™, ResLink™, ResTraxx™ 1 an Appareils de traitement ResMed 2 ans Accessoires, réservoirs et masques (y compris entourage rigide, bulle, harnais et circuit respiratoire). Les appareils à usage unique sont exclus. 90 jours Remarque : certains de ces modèles ne sont pas disponibles dans tous les pays. En cas de défaillance du produit dans des conditions normales d’utilisation, ResMed, à son entière discrétion, répare ou remplace le produit défectueux ou toute pièce. Cette garantie limitée ne couvre pas : a) tout dommage résultant d’une utilisation incorrecte, d’un usage abusif ou d’une modification ou transformation opérée sur le produit ; b) les réparations et l’entretien effectués par tout service de réparation sans l’autorisation expresse de ResMed ; c) tout dommage ou contamination causé par de la fumée de cigarette, de pipe, de cigare ou autre ; d) tout dommage causé par de l’eau renversée sur ou dans un appareil. La garantie est annulée pour les produits vendus ou revendus dans un pays autre que celui où ils ont été achetés à l’origine. Les réclamations au titre de la garantie pour les produits défectueux doivent être présentées au lieu d’achat par le client initial. Cette garantie remplace toute autre garantie expresse ou implicite, y compris toute garantie implicite de qualité marchande ou d’adéquation à un usage particulier. Certains pays ou états n’autorisent pas les limitations de durée pour les garanties implicites ; il est donc possible que la limitation susmentionnée ne s’applique pas à votre cas. 35 Français Réparations ResMed ne peut être tenue pour responsable de tout dommage accessoire ou indirect résultant de la vente, de l’installation ou de l’utilisation de tout produit ResMed. Certains pays ou états n’autorisent ni l’exclusion ni la limitation des dommages accessoires ou indirects ; il est donc possible que la limitation susmentionnée ne s’applique pas à votre cas. Cette garantie vous octroie des droits reconnus par la loi, et vous pouvez également avoir d’autres droits qui varient en fonction du pays où vous habitez. Pour de plus amples informations sur vos droits de garantie, veuillez contacter votre prestataire de santé ou ResMed. Guide et déclaration du fabricant – Émissions et immunité électromagnétiques Guide et déclaration du fabricant – Émissions électromagnétiques Le C-Series Tango est prévu pour une utilisation dans l’environnement électromagnétique décrit ci-dessous. Le client ou l’utilisateur du C-Series Tango doit s’assurer que celui-ci est utilisé dans un environnement de ce type. Essai de contrôle des émissions Conformité Environnement électromagnétique – Guide Émissions RF CISPR11 Groupe 1 Le C-Series Tango utilise de l’énergie RF uniquement pour son fonctionnement interne. Par conséquent, les émissions RF sont très faibles et ne devraient pas causer d’interférences avec l’équipement électronique environnant. Émissions RF CISPR 11 Classe B Rayonnements harmoniques CEI 61000-3-2 Classe A Le C-Series Tango peut être utilisé dans tous les établissements, y compris les établissements domestiques et ceux directement raccordés au réseau public de distribution à basse tension qui fournit de l’électricité aux bâtiments utilisés à des fins domestiques. Émissions dues aux fluctuations de tension/au papillotement CEI 61000-3-3 Conforme Des précautions particulières doivent être prises avec les appareils électromédicaux quant à la compatibilité électromagnétique, et leur installation et leur mise en service doivent être effectuées conformément aux informations sur la compatibilité électromagnétique fournies dans ce document. Avertissements : le C-Series Tango ne doit pas être utilisé à proximité d’autres équipements ni posé sur ou sous d’autres équipements. S’il est impossible de l’utiliser autrement, vous devez vous assurer qu’il fonctionne normalement lorsqu’il est placé à l’endroit où il sera utilisé. Il est déconseillé d’utiliser des accessoires (par exemple des humidificateurs) autres que ceux spécifiés dans ce manuel. Ils peuvent entraîner une augmentation des émissions ou une réduction de l’immunité du C-Series Tango. 36 Guide et déclaration du fabricant – Immunité électromagnétique Essai de contrôle de l’immunité Niveau d’essai CEI 60601-1-2 Niveau de conformité Environnement électromagnétique – Guide Décharge électrostatique CEI 61000-4-2 Contact ± 6 kV Contact ± 6 kV Air ± 8 kV Air ± 8 kV Les sols doivent être en bois, en ciment ou carrelés. Si le revêtement des sols est synthétique, l’humidité relative doit être d’au moins 30 %. Transitoires/salves électriques rapides CEI 61000-4-4 ± 2 kV pour les lignes d’alimentation ± 2 kV ± 1 kV pour les lignes d’alimentation/de signalisation Sans objet ± 1 kV en mode différentiel ± 1 kV en mode différentiel ± 2 kV en mode commun ± 2 kV en mode commun < 5 % Ut (> 95 % de baisse en Ut) pour 0,5 cycle < 12 V (> 95 % de baisse en 240 V) pour 0,5 cycle 40 % Ut (60 % de baisse en Ut) pour 5 cycles 96 V (60 % de baisse en 240 V) pour 5 cycles 70 % Ut (30 % de baisse en Ut) pour 25 cycles 168 V (30 % de baisse en 240 V) pour 25 cycles < 5 % Ut (> 95 % de baisse en Ut) pour 5 sec < 12 V (> 95 % de baisse en 240 V) pour 5 sec 3 A/m 3 A/m Surtension CEI 61000-4-5 Baisses de tension, brèves coupures de courant et variations de tension sur les lignes d’alimentation CEI 61000-4-11 Champ magnétique de la fréquence d’alimentation (50/60 Hz) CEI 61000-4-8 Remarque : Ut correspond à la tension secteur AC avant l’application du niveau d’essai. 37 La qualité de l’alimentation secteur doit être celle d’un environnement commercial ou médical standard. La qualité de l’alimentation secteur doit être celle d’un environnement commercial ou médical standard. La qualité de l’alimentation secteur doit être celle d’un environnement commercial ou médical standard. Si l’utilisateur du C-Series Tango ne peut pas supporter l’interruption du traitement en cas de coupure de courant, il est recommandé d’utiliser le C-Series Tango avec une source d’alimentation ininterruptible. Les champs magnétiques de la fréquence réseau doivent correspondre à ceux d’un environnement commercial ou médical standard. Français Le C-Series Tango est prévu pour une utilisation dans l’environnement électromagnétique décrit ci-dessous. Le client ou l’utilisateur du C-Series Tango doit s’assurer que celui-ci est utilisé dans un environnement de ce type. Guide et déclaration du fabricant – Immunité électromagnétique Le C-Series Tango est prévu pour une utilisation dans l’environnement électromagnétique décrit ci-dessous. Le client ou l’utilisateur du C-Series Tango doit s’assurer que celui-ci est utilisé dans un environnement de ce type. Essai de contrôle de l’immunité CEI60601-1-2 Niveau d’essai Niveau de conformité Environnement électromagnétique – Guide Les équipements de communication RF portables et mobiles ne doivent pas être utilisés à proximité des composants du C-Series Tango, y compris les câbles, à une distance inférieure à la distance de séparation recommandée calculée sur la base de l’équation applicable à la fréquence de l’émetteur. Distance de séparation recommandée RF transmises par conduction CEI 61000-4-6 3 Vrms 150 kHz à 80 MHz 3 Vrms d = 1,17 √P RF transmises par radiation CEI 61000-4-3 10 V/m 80 MHz à 2,5 GHz 10 V/m d = 0,35 √P 80 MHz à 800 MHz d = 0,70 √P 800 MHz à 2,5 GHz où P correspond à la puissance nominale de sortie maximale de l’émetteur en watts (W) indiquée par le fabricant de l’émetteur et d correspond à la distance de séparation recommandée en mètres (m). L’intensité des champs d’émetteurs RF fixes, telle qu’elle est déterminée par l’étude electromagnétique du sitea doit être inférieure au niveau de conformité pour chaque plage de fréquences.b Des interférences peuvent se produire à proximité d’équipements portant le symbole suivant : Remarque 1 : à 80 MHz et 800 MHz, la plage de fréquences supérieure s’applique. Remarque 2 : il est possible que ces directives ne s’appliquent pas à toutes les situations. La propagation électromagnétique dépend de l’absorption et de la réflexion des structures, des objets et des personnes. L’intensité des champs d’émetteurs fixes, tels que des stations de base pour des radiotéléphones (cellulaires/ sans fil) et des radios mobiles, la radio amateur, les radios AM et FM et la télévision, ne peut pas être prédite théoriquement avec précision. Pour évaluer l’environnement électromagnétique dû aux émetteurs RF fixes, une étude électromagnétique du site doit être envisagée. Si l’intensité du champ mesurée à l’endroit où le C-Series Tango est utilisé dépasse le niveau de conformité RF applicable figurant ci-dessus, il faudra s’assurer du bon fonctionnement du C-Series Tango dans un tel environnement. En cas d’anomalie, des mesures supplémentaires devront peut-être être prises, par exemple changement de position ou d’endroit pour le C-Series Tango. a b Pour la plage de fréquences 150 kHz à 80 MHz, l’intensité des champs doit être inférieure à 10 V/m. 38 Distances de séparation recommandées entre les équipements de communication RF portables et mobiles et le C-Series Tango. Le C-Series Tango est prévu pour une utilisation dans un environnement dans lequel les perturbations radioélectriques par radiation sont contrôlées. Le client ou l’utilisateur du C-Series Tango peut contribuer à la prévention des perturbations électromagnétiques en respectant les distances minimales recommandées ci-dessous pour la séparation entre les équipements de communication RF portables et mobiles (émetteurs) et le C-Series Tango, en fonction de la puissance maximale de sortie de l’équipement de communication. 150 kHz à 80 MHz 80 MHz à 800 MHz 800 MHz à 2,5 GHz d = 1,17 √P d = 0,35 √P d = 0,7 √P 0,01 0,17 0,04 0,07 0,1 0,37 0,11 0,22 1 1,17 0,35 0,7 10 3,69 1,11 2,21 100 11,70 3,50 7,0 Pour les émetteurs avec une puissance nominale de sortie maximale ne figurant pas dans le tableau ci-dessus, la distance de séparation recommandée d en mètres (m) peut être déterminée à l’aide de l’équation applicable à la fréquence de l’émetteur, où P correspond à la puissance nominale de sortie maximale de l’émetteur en watts (W) indiquée par le fabricant de l’émetteur. Remarque 1 : à 80 MHz et 800 MHz, la distance de séparation pour la plage de fréquences supérieure s’applique. Remarque 2 : il est possible que ces directives ne s’appliquent pas à toutes les situations. La propagation électromagnétique dépend de l’absorption et de la réflexion des structures, des objets et des personnes. 39 Français Distance de séparation en fonction de la fréquence de l’émetteur (m) Puissance nominale de sortie maximale de l’émetteur W 40 ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA ResMed (UK) Ltd (EU Authorized Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK ResMed Offices Australia, Austria, Brazil, China, Finland, France, Germany, Hong Kong, India, Japan, Malaysia, Netherlands, New Zealand, Norway, Singapore, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com for contact details). C-Series Tango Protected by patents: US 5199424, US 5522382, US 6705315. Protected by design registrations: AU 307932, AU 307933, AU 307934, AU 307935, AU 309812, AU 310045, AU 310046, AU 310165, AU 311258, AU 311273, AU 311486, AU 311487, CN 200630002263.2, CN 200630002264.7, CN 200630002265.1, CN 200630002266.6, CN 200630002267.0, CN 200630002269.X, CN 200630002270.2, CN 200630002271.7, CN 200630137884.1, EU 599634, EU 600176, JP 1293705, JP 1294379, JP 1294380, JP 1296487, JP 1296603, JP 1296604, JP 1296605, JP 1303495, NZ 407118, NZ 408195, NZ 408705, NZ 408706, NZ 408707, NZ 408708, NZ 408709, NZ 408710, US D542900, US D544591, US D545954, US D549321, US D549810, US D550349. Patents and designs pending. Tango, Mirage, Mirage Activa, Mirage Vista, Ultra Mirage, Mirage Swift, Mirage Liberty, Mirage Quattro and Meridian are trademarks of ResMed Ltd. Mirage, Mirage Activa, Mirage Vista, Ultra Mirage, Mirage Swift and Meridian are registered in US Patent and Trademark Office. Papillon and Silent Papillon are trademarks of MAP Medizin-Technologie GmbH. © 2007 ResMed Ltd. 34835/1 07 10 C-Series Tango user AMER ML Global leaders in sleep and respiratory medicine   www.resmed.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44

ResMed Tango Manual de usuario

Tipo
Manual de usuario
Este manual también es adecuado para

En otros idiomas