Greenlee DM-110 El manual del propietario

Categoría
Medir, probar
Tipo
El manual del propietario

Este manual también es adecuado para

999 3788.3 © 2000 Greenlee Textron IM 1463 REV 1 12/00
DM-100 • DM-110
DIGITAL
MULTIMETERS
MULTÍMETROS
DIGITALES
MULTIMETRES
NUMERIQUES
INSTRUCTION MANUAL
INSTRUCTION MANUAL
MANUAL DE INSTRUCCIONES
MANUEL D’INSTRUCTIONS
Read and understand all of the instructions and safety
information in this manual before operating or servicing this tool.
Lea y entienda todas las instrucciones y la información sobre
seguridad que aparecen en este manual, antes de manejar estas
herramientas o darles mantenimiento.
Lire attentivement et bien comprendre toutes les instructions et
les informations sur la sécurité de ce manuel avant d‘utiliser ou de
procéder à l‘entretien de cet outil.
2
Description
The Greenlee DM-100 and DM-110 Digital Multimeters are hand-held testing devices with the
following measurement capabilities: AC and DC voltage, AC and DC current, and resistance.
They also check diodes and verify continuity. In addition, the DM-110 measures frequency and
capacitance. The DM-110 is a true RMS reading meter.
Both models feature automatic power off to conserve the batteries and a bar graph display
which is more sensitive than the numeric display—useful for detecting faulty contacts,
potentiometer clicks, and signal spikes.
Safety
Safety is essential in the use and maintenance of Greenlee tools and equipment. This
instruction manual and any markings on the tool provide information for avoiding hazards and
unsafe practices related to the use of this tool. Observe all of the safety information provided.
Purpose
This instruction manual is intended to familiarize all personnel with the safe operation and
maintenance procedures for Greenlee DM-100 and DM-110 Digital Multimeters.
Keep this manual available to all personnel.
Replacement manuals are available upon request at no charge.
Greenlee and are registered trademarks of Greenlee Textron.
KEEP THIS MANUAL
DM-100 DM-110
3
Important Safety Information
Read and understand this material before operating or servicing
this equipment. Failure to understand how to safely operate this
tool can result in an accident causing serious injury or death.
This symbol is used to call your attention to hazards or unsafe practices which could
result in an injury or property damage. The signal word, defined below, indicates the
severity of the hazard. The message after the signal word provides information for
preventing or avoiding the hazard.
SAFETY ALERT SYMBOL
Immediate hazards which, if not avoided, WILL result in severe injury or death.
Hazards which, if not avoided, COULD result in severe injury or death.
Hazards or unsafe practices which, if not avoided, MAY result in injury or property
damage.
4
Important Safety Information
Electric shock and fire hazard:
Do not expose this unit to rain or moisture.
Do not use the unit if it is wet or damaged.
Use test leads or accessories that are appropriate for the application. See the
category and voltage rating of the test lead or accessory.
Inspect the test leads or accessory before use. They must be clean and dry, and the
insulation must be in good condition.
Use this unit for the manufacturer’s intended purpose only, as described in this
manual. Any other use can impair the protection provided by the unit.
Failure to observe these warnings can result in severe injury or death.
Do not operate with the case open.
Before opening the case, remove the test leads from the circuit and shut off the unit.
Failure to observe these warnings can result in severe injury or death.
Electric shock hazard:
Contact with live circuits can result in severe injury or death.
Electric shock hazard:
Do not apply more than the rated voltage between any two input terminals, or
between any input terminal and earth ground.
Do not contact the test lead tips or any uninsulated portion of the accessory.
Failure to observe these warnings can result in severe injury or death.
DM-100 DM-110
5
Important Safety Information
Do not attempt to repair this unit. It contains no user-serviceable parts.
Do not expose the unit to extremes in temperature or high humidity.
See Specifications.
Failure to observe these precautions can result in injury and can damage the unit.
Do not change the measurement function while the test leads are connected to a
component or circuit.
Failure to observe this precaution can result in injury and can damage the unit.
The fuse is an integral part of the overvoltage protection. When fuse replacement is
necessary, see Specifications for the correct type, size, and capacity. Using any other
type of fuse will void the overvoltage protection rating of the unit.
Failure to observe this warning can result in severe injury or death.
Unless measuring voltage, current, or frequency, shut off and lock out power.
Make sure that all capacitors are discharged. Voltage must not be present.
Set the selector and connect the test leads so that they correspond to the intended
measurement. Incorrect settings or connections can result in a blown fuse.
Using this unit near equipment that generates electromagnetic interference can
result in unstable or inaccurate readings.
6
Identification
1. Display 4-digit LCD and bar graph
2. Function Buttons See explanations under Using the Features
3. Selector Selects a function or turns power OFF
4. Hz Positive input terminal for all measurements
except
10 A current
VΩµA( Hz on DM-110 only)
5. COM Negative, common or ground input terminal for all measurements
6. 10A Positive input terminal for current measurements up to 10 A
1
2
5
6
3
4
Symbols on the Unit
WarningRead the instruction manual
Battery
Fuse
Double insulation
DM-100 DM-110
7
Display Icons
Both Models
7. Manual range is selected
8. DC measurement is selected
9. AC measurement is selected
10. Low battery
11. APO Auto Power Off is enabled
12. Diode
13. Continuity is selected
14. µ Micro (10
-6
)
15. m Milli (10
-3
)
16. V Volts
17. A Amps
18. M Mega (10
6
)
19. k Kilo (10
3
)
20. Ohms
21. Hz Hertz (frequency
or cycles per
second)
7
9
8
10
22
14,15,16,17
18,19,20,21
12 1311
7
8910
1123 12
13
14,15,16,17
24,25,26,27
18,19,20,21
DM-100 Only
22. Hold function is enabled
DM-110 Only
23. Hold function is enabled
24. n nano (10
-9
)
25. µ Micro (10
-6
)
26. m Milli (10
-3
)
27. F Farads
Note: Unidentified icons are not used
on these models.
DM-100 display
DM-110 display
8
AC Measurement
AC measurements are usually displayed as RMS (root-mean-square) values. Two AC
measurement methods are average-responding RMS calibrated and true RMS-reading.
The average-responding RMS calibrated method takes the average value of the input signal,
multiplies it by 1.11, and displays the result. This method is accurate if the input signal is a
pure sine wave. The Greenlee DM-100 is an average-responding meter.
The true RMS-reading method uses internal circuitry to read the true RMS value. This method
is accurate, within the specified crest factor limitations, whether the input signal is a pure sine
wave, a square wave, triangle wave, half wave or signal with harmonics. The ability to read
true RMS provides much more measurement versatility. The Greenlee DM-110 is a true RMS
meter.
The Waveforms and Crest Factors table shows some typical AC signals and their RMS values.
Waveforms and Crest Factors
*The crest factor is the ratio of the peak value to the RMS value; it is represented by the
Greek letter ξ.
Waveform
RMS Value 100 100 100 100
Average Value 90 100 87 64
Crest Factor*
1.414 1 1.73 2
(ξ)
DM-100 DM-110
9
Button Press momentarily to: Notes:
Blue select AC or DC when
(no label) measuring current or voltage
select secondary functions
RANGE change from autoranging Press and hold to change from
to manual ranging manual ranging to autoranging.
select ranges when using
the manual ranging mode
select trigger levels when
measuring frequency
PWR RST
turn power back on after auto
(DM-100 only)
power off
Hz view the frequency while While in the Hz mode, the RANGE
(DM-110
only)
measuring AC voltage or button does not change the
current frequency range. However, it changes
the voltage range.
HOLD hold a measured value for all Conversions are made but the display
functions. Press again to does not update.
return to normal mode
Using the Features
10
Operation
1. See the Settings Table. Set the selector to the proper setting, press the unlabeled blue
button to select the desired measurement, and connect the test leads to the meter. Start
with the highest measurement setting.
2. See Typical Measurements for specific measurement instructions.
3. Test the unit on a known functioning circuit or component.
If the unit does not function as expected on a known functioning circuit, replace the
battery and/or fuse.
If the unit still does not function as expected, send the unit to Greenlee for repair.
See the address shown under Warranty.
4. Take the reading from the circuit or component to be tested. If the resolution is not
satisfactory, remove the meter from the circuit and change to the next lower range if using
manual ranging.
DM-100 DM-110
11
Settings Table
set the selector
this icon will connect and connect
To measure this …
to this symbol …
appear on the red the black
the display … lead to … lead to …
Capacitance* F Hz COM
(DM-110 only) VΩµA
Continuity** and press the BLUE VΩµA COM
button one time
µA
Current (AC) and press the BLUE µAVΩµA COM
button one time
Current (DC) µA VΩµA COM
µA
Current (AC) 10A
(10 A max) and press the BLUE 10 A COM
button one time A
Current (DC) 10A 10 A COM
(10 A max) A
Diode*** and press the BLUE VVΩµA COM
button two times
Frequency Hz kHz or MHz Hz COM
(DM-110 only) VΩµA
Resistance M or k VΩµA COM
V
Voltage (AC) and press the BLUE mV or V VΩµA COM
button one time
Voltage (DC) V VΩµA COM
mV or V
* Discharge capacitor before measurement.
** Tone sounds if the measured resistance is less than approximately 30 .
*** Diode voltage is shown if conducting, or OL is displayed for non-conductance.
3200
µA max for DM-100, 4000 µA max for DM-110.
Operation (contd)
12
Operation (contd)
Auto Power Off
The meter defaults to automatically turn itself off after a period of inactivity from the user.
on the display indicates that the feature is enabled. The meter beeps approximately 15
seconds prior to turning off. The DM-100 turns off after 10 minutes, and the DM-110 turns off
after 30 minutes. The present reading is held in memory. To turn the meter back on, press any
button or turn the selector. The last reading will be displayed in the Hold mode. Press HOLD to
return to normal operation.
Disabling Auto Power Off
To disable the Auto Power Off feature, press and hold the RANGE button while turning the
meter on.
Note: When the meter is turned off and back on with the selector, the meter returns to the default.
However, the last value is not held in memory.
DM-100 DM-110
13
Typical Measurements
Voltage Measurement
Current Measurement
14
Typical Measurements (contd)
Resistance Measurement
Continuity Check
DM-100 DM-110
15
Typical Measurements (contd)
Capacitance Measurement
Diode Measurement
Forward Bias
Reverse Bias
16
Accuracy: DM-100
See the Specifications section for operating conditions and temperature coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount)
at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 80% relative humidity.
For the DM-100, AC accuracy is specified for sine waves only.
* In autorange mode, ranges change at approximately 3200 counts.
In manual range mode, range extends to 3400 counts.
AC Voltage
Range* Accuracy Frequency Range Input Impedance
3.200 V ± (1.50% + 0.005 V)
32.00 V ± (1.50% + 0.05 V)
320.0 V ± (1.50% + 0.5 V)
750 V ± (1.50% + 5 V) inal
40 to 500 Hz 10 M, 100 pF
DC Current
Range* Accuracy Burden Voltage
320.0 µA ± (1.7% + 0.2 µA) 1 V max
3200 µA ± (1.7% + 2 µA) 10 V max
10.00 A ± (2% + 0.02 A) 2 V max
AC Current
Range* Accuracy Burden Voltage
320.0 µA ± (2.2% + 0.5 µA) 1 V max
3200 µA ± (2.2% + 5 µA) 10 V max
10.00 A ± (2.5% + 0.05 A) 2 V max
DM-100 DM-110
17
Accuracy: DM-100 (contd)
Resistance
Range* Accuracy
320.0 Ω± (1% + 0.4 )
3.200 kΩ± (0.75% + 0.003 k)
32.00 kΩ± (0.75% + 0.03 k)
320.0 kΩ± (0.75% + 0.3 k)
3.200 MΩ± (1% + 0.003 M)
32.00 MΩ± (2% + 0.05 M)
Diode Test
Test Current (Typical): 1.5 mA
Open Circuit Voltage: Less than 3.1 VDC
Continuity
Threshold: Tone sounds if the measured resistance is less than approximately 30 .
Response Time: Approximately 100 ms
Typical open circuit voltage: 1.5V
* In autorange mode, ranges change at approximately 3200 counts.
In manual range mode, range extends to 3400 counts.
DC Voltage
Range* Accuracy Input Impedance
320.0 mV ± (0.5% + 0.2 mV) over 1000 M
3.200 V ± (0.5% + 0.002 V)
32.00 V ± (0.5% + 0.02 V)
320.0 V ± (0.5% + 0.2 V)
1000 V ± (0.5% + 2 V)
10 M
18
Accuracy: DM-110
See the Specifications section for operating conditions and temperature coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount)
at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 80% relative humidity.
For the DM-110, AC accuracy specification applies for sine waves at full scale and non-sine
waves below half scale.
For crest factors from 1.4 to 3, add 1.5% to accuracy.
For crest factors from 3 to 4, add 3% to accuracy.
DC Current
Range Accuracy Burden Voltage
400.0 µA ± (1.7% + 0.2 µA) 1 V max
4000 µA ± (1.7% + 2 µA) 10 V max
10.00 A ± (2% + 0.02 A) 2 V max
AC Voltage
Range Accuracy Frequency Range Input Impedance
400.0 mV unspecified
4.000 V ± (1.5% + 0.005 V)
40.00 V ± (1.3% + 0.05 V) 40 to 500 Hz 10 M, 100 pF
400.0 V ± (1.3% + 0.5 V)
750 V ± (1.3% + 5 V)
AC Current
Range Accuracy Burden Voltage
400.0 µA ± (2.2% + 0.5 µA) 1 V max
4000 µA ± (2.2% + 5 µA) 10 V max
10.00 A * ± (2.5% + 0.05 A) 2 V max
* 10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off)
DM-100 DM-110
19
Accuracy: DM-110 (contd)
Accuracies are for film capacitors (capacitors with negligible dielectric absorption)
*Accuracy specified for values less than 1/2 full scale. May have ripple within accuracy spec limits.
Typical open circuit voltage: 1.3V
Resistance
Range Accuracy
400.0 Ω± (1.0% + 0.3 )
4.000 kΩ± (0.75% + 0.002 k)
40.00 kΩ± (0.75% + 0.02 k)
400.0 kΩ± (0.75% + 0.2 k)
4.000 MΩ± (1.0% + 0.003 M)
40.00 MΩ± (1.5% + 0.05 M)
Capacitance
Range Accuracy
4.000 nF ± (3.0% + 0.010 nF)
40.00 nF ± (3.0% + 0.10 nF)
400.0 nF ± (2.0% + 0.8 nF)
4.000 µF ± (2.0% + 0.008 µF)
40.00 µF ± (2.0% + 0.08 µF)
400.0 µF ± (2.0% + 0.8 µF)
4.000 mF* ± (5.0% + 0.020 mF)
40.00 mF* ± (5.0% + 0.20 mF)
DC Voltage
Range Accuracy Input Impedance
400.0 mV ± (0.5% + 0.2 mV) over 1000 M
4.000 V ± (0.5% + 0.002 V)
40.00 V ± (0.5% + 0.02 V)
400.0 V ± (0.5% + 0.2 V)
1000 V ± (0.5% + 2 V)
10 M
20
Accuracy: DM-110 (contd)
Diode Test
Test Current (Typical): 1.5 mA
Open Circuit Voltage: Less than 3.1 VDC
Continuity
Threshold: Tone sounds if the measured resistance is less than approximately 30 .
Response Time: Approximately 100 ms
*Sensitivity is 1.5 V rms below 20 Hz
The frequency measuring function has four selectable trigger levels. Use the range button
to select.
Frequency
Range Accuracy Sensitivity
4.000 kHz ± (0.01% + 0.001 kHz)
40.00 kHz ± (0.01% + 0.01 kHz) 100 mV rms *
400.0 kHz ± (0.01% + 0.1 kHz)
4.000 MHz ± (0.01% + 0.001 MHz) 250 mV rms
40.00 MHz ± (0.01% + 0.01 MHz) 1 V rms
DM-100 DM-110
21
Specifications
Display:
DM-100: LCD (3200) and 70-segment bar graph
In autorange mode, ranges change at approximately 3200 counts. In manual range
mode, range extends to 3400 counts.
DM-110: LCD (4000) and 82-segment bar graph
Polarity: Automatic
Sampling Rate:
Numeric Display: 2 per second
Bar Graph Display: 12 per second
Temperature Coefficient: 0.15 x (Accuracy) per °C below 18 °C or above 28 °C
Automatic Power-Off:
DM-100: After 10 minutes of inactivity
DM-110: After 30 minutes of inactivity
Overload Protections:
Volts: See Overvoltage Protection Categories
Amps: A: 16A/500V HBC F Fuse, Interrupting Rating 10 kA, 1/4 x 1-1/4"
Other Functions: 600 VDC/VAC RMS
Overvoltage Protection Categories: VΩµA Terminal:
Category III, 600 VAC and 600 VDC
Category II, 1000 VAC and 1000 VDC
Operating Conditions:
0 °C to 30 °C (32 °F to 86 °F), 0% to 80% Relative Humidity
30 °C to 40 °C (86 °F to 104 °F), 0% to 75% Relative Humidity
40 °C to 50 °C (104 °F to 122 °F), 0% to 45% Relative Humidity
Indoor use only.
Storage Conditions:
-20 °C to 60 °C (-4 °F to 140 °F), 0% to 80% Relative Humidity
Remove batteries
Elevation: 2000 m (6500') maximum
Pollution Degree: 2
Battery:
DM-100: Two AAA 1.5-Volt batteries (NEDA 24A or IEC LR03)
DM-110: One 9-Volt battery (NEDA 1604 or IEC 6F22)
Battery Life:
DM-100: Approximately 500 hours with alkaline
DM-110: Approximately 300 hours with alkaline
22
Overvoltage Installation Categories
These definitions were derived from the international safety standard for insulation
coordination as it applies to measurement, control, and laboratory equipment. These
overvoltage categories are explained in more detail by the International Electrotechnical
Commission; refer to either of their publications: IEC 1010-1 or IEC 60664.
Overvoltage Category I
Signal level. Electronic and telecommunication equipment, or parts thereof. Some examples
include transient-protected electronic circuits inside photocopiers and modems.
Overvoltage Category II
Local level. Appliances, portable equipment, and the circuits they are plugged into. Some
examples include light fixtures, televisions, and long branch circuits.
Overvoltage Category III
Distribution level. Permanently installed machines and the circuits they are hard-wired to.
Some examples include conveyor systems and the main circuit breaker panels of a buildings
electrical system.
Overvoltage Category IV
Primary supply level. Overhead lines and other cable systems. Some examples include cables,
meters, transformers, and other exterior equipment owned by the power utility.
DM-100 DM-110
23
Maintenance
1. Disconnect the unit from the circuit. Turn the unit OFF.
2. Remove the screws from the back cover.
3. Remove the back cover.
4. Lift battery holder out of back cover.
5. Replace the batteries (observe polarity) and/or fuse.
6. Replace the cover, screws and stand.
Cleaning
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or
solvents.
Do not attempt to repair this unit. It contains no user-serviceable parts.
Do not expose the unit to extremes in temperature or high humidity.
See Specifications.
Failure to observe these precautions can result in injury and can damage the unit.
The fuse is an integral part of the overvoltage protection. When fuse replacement is
necessary, see Specifications for the correct type, size, and capacity. Using any other
type of fuse will void the overvoltage protection rating of the unit.
Failure to observe this warning can result in severe injury or death.
Battery and Fuse Replacement
Before opening the case, remove the test leads from the circuit and shut off the unit.
Failure to observe this warning can result in severe injury or death.
24
Greenlee Textron / Subsidiary of Textron Inc. 4455 Boeing Dr., Rockford, IL 61109-2988 815/397-7070
DM-100 • DM-110
25
Descripción
Los Multímetros digitales modelos DM-100 y DM-110 de Greenlee son instrumentos de verificación
capaces de efectuar los siguientes tipos de mediciones: tensión alterna y continua, corriente alterna
y continua, y resistencia. Estas unidades son de bolsillo y caben perfectamente en la palma de la
mano. También sirven para verificar diodos y continuidad. Además, el modelo DM-110 es capaz
de efectuar mediciones de frecuencia y capacitancia. El modelo DM-110 es un multímetro de
lectura de valores eficaces reales.
Ambos modelos incluyen la función de apagado automático para conservar la vida útil de las pilas,
además de un gráfico de barras, el cual es más sensible que la pantalla numérica – útil para detectar
fallas en contatos, potenciómetros y picos de señal.
Acerca de la seguridad
Es fundamental observar métodos seguros al utilizar y dar mantenimiento a las herramientas y
equipo Greenlee. Este manual de instrucciones y todas las marcas que ostenta la herramienta le
ofrecen la información necesaria para evitar riesgos y hábitos poco seguros relacionados con su
uso. Siga toda la información sobre seguridad que se proporciona.
Propósito de este manual
Este manual de instrucciones tiene como propósito familiarizar a todo el personal con los
procedimientos de operación y mantenimiento seguros para los Multímetros digitales
modelos DM-100 y DM-110 de Greenlee.
Manténgalo siempre al alcance de todo el personal.
Puede obtener copias adicionales de manera gratuita, previa solicitud.
Greenlee y son marcas registradas de Greenlee Textron.
CONSERVE ESTE MANUAL
26
Importante Información sobre Seguridad
Lea y entienda este documento antes de manejar esta herramienta o
darle mantenimiento. Utilizarla sin comprender cómo manejarla de
manera segura podría ocasionar un accidente y, como resultado
de éste, graves lesiones o incluso la muerte.
Este símbolo se utiliza para indicar un riesgo o práctica poco segura que podría ocasionar
lesiones o daños materiales. Cada uno de los siguientes términos denota la gravedad del
riesgo. El mensaje que sigue a dichos términos le indica cómo puede evitar o prevenir ese
riesgo.
SÍMBOLO DE ALERTA
SOBRE SEGURIDAD
Peligros inmediatos que, de no evitarse, OCASIONARÁN graves lesiones o incluso
la muerte.
Peligros que, de no evitarse, PODRÍAN OCASIONAR graves lesiones o incluso la muerte.
Peligro o prácticas peligrosas que, de no evitarse, PUEDEN OCASIONAR lesiones
o daños materiales.
DM-100 • DM-110
27
Importante Información sobre Seguridad
Peligro de electrocución e incendio:
No exponga esta unidad ni a la lluvia ni a la humedad.
No utilice esta unidad si se encuentra mojada o dañada.
Utilice cables de prueba y accesorios que sean apropiados para la aplicación que se va a
realizar. Consulte la información sobre categoría y voltaje nominal del cable de prueba o
el accesorio.
Revise minuciosamente los cables de prueba o el accesorio, antes de utilizarlos. Deberán
estar limpios y secos, y su forro aislante deberá hallarse en buenas condiciones.
Utilícela únicamente para el propósito para el que ha sido diseñada por el fabricante,
tal como se describe en este manual. Cualquier otro uso puede menoscabar la protección
proporcionada por la unidad.
De no observarse estas advertencias pueden sufrirse graves lesiones o incluso la muerte.
No haga funcionar esta unidad con la caja abierta.
Antes de abrir la caja, retire del circuito los cables de prueba y apague la unidad.
De no observarse estas advertencias pueden sufrirse graves lesiones o incluso la muerte.
Peligro de electrocución:
El contacto con circuitos activados puede ocasionar graves lesiones o
incluso la muerte.
Peligro de electrocución:
No aplique más del voltaje nominal entre dos terminales de entrada cualesquiera,
o entre una terminal de entrada cualquiera y una conexión a tierra.
No toque las puntas de los cables de prueba ni ninguna parte del accesorio que
carezca de forro aislante.
De no observarse estas advertencias pueden sufrirse graves lesiones o incluso la muerte.
28
Importante Información sobre Seguridad
No intente reparar esta unidad, ya que contiene partes que deben recibir mantenimiento
por parte de un profesional.
No exponga la unidad a ambientes de temperatura extrema o altos niveles de humedad;
Consulte las Especificaciones.
De no observarse estas precauciones podrían sufrirse lesiones o daños a la unidad.
No cambie la función de medición mientras los cables de prueba estén conectados a un
componente o circuito.
De no observarse esta advertencia podrían sufrirse lesiones o daños a la unidad.
El fusible es una parte integral para la protección contra sobretensión. Cuando sea necesario
reemplazarlo, consulte la sección “Especificaciones” para saber qué tipo, tamaño y
capacidad debe tener. Utilizar cualquier otro tipo de fusible anulará la clasificación de
protección de sobretensión de la unidad.
De no observarse esta advertencia pueden sufrirse graves lesiones o incluso la muerte.
A menos que vaya a medir tensión, corriente o frecuencia, apague y bloquee la energía.
Asegúrese de que todos los condensadores estén totalmente sin carga. No debe haber
tensión alguna.
Coloque el interruptor de selección y conecte los cables de prueba de modo que
correspondan al tipo de medición que se desea efectuar. Si se colocan o se conectan
incorrectamente puede quemarse un fusible.
Al utilizar esta unidad cerca de equipo que genere interferencia electromagnética quizá
se obtenga una lectura inexacta e inestable.
DM-100 • DM-110
29
Identificación
1. Pantalla Pantalla de cristal líquido (LCD) de 4 dígitos y gráfico de barras
2. Botones Selectores Véase la explicación pertinente en la sección “Cómo utilizar las
de funciones distintas funciones”
3. Selector Selecciona una función o apaga (OFF) la unidad
4. Hz Terminal de señal de entrada positiva para todas las mediciones
VΩµA
excepto
corriente de 10A ( Hz en modelo DM-110 únicamente)
5. COM Terminal de entrada a tierra, común o negativa, para todo tipo de
mediciones
6. 10A Terminal de entrada positiva para mediciones de corriente de hasta
10 A
1
2
5
6
3
4
Símbolos en la unidad
AdvertenciaLea el manual de instrucciones
Pila
Fusible
Doble forro aislante
30
Iconos de la pantalla
Ambos modelos
7. Se selecciona Medición manual de escala
8. Se selecciona medición de CC
9. Se selecciona medición de CA
10. Pila baja
11. APO Se activa la función Apagado automático
12. Diodo
13. Se selecciona Continuidad
14. µ Micro (10
-6
)
15. m Mili (10
-3
)
16. V Voltios
17. A Amperios
18. M Mega (10
6
)
19. k Kilo (10
3
)
20. Ohmios
21. Hz Herzios
(frecuencia o ciclos
por segundo)
7
9
8
10
22
14,15,16,17
18,19,20,21
12 1311
7
8910
1123 12
13
14,15,16,17
24,25,26,27
18,19,20,21
Modelo DM-100 únicamente
22. Se activa la función Hold
(Retención de datos en pantalla)
Modelo DM-110 únicamente
23. Se activa la función Hold
(Retención de datos en pantalla)
24. n nano (10
-9
)
25. µ Micro (10
-6
)
26. m Mili (10
-3
)
27. F Faradios
Nota: Los iconos sin identificar no
se utilizan en estos modelos.
Pantalla del modelo DM-100
Pantalla del modelo DM-110
DM-100 • DM-110
31
Medición de corriente alterna
Las mediciones de corriente alterna generalmente se muestran como valores eficaces (RMS o root
mean squared). Existen dos métodos de medición de corriente alterna: calibrados para responder
al valor eficaz medio y a una lectura de valores eficaces reales.
El método calibrado para responder al valor eficaz medio toma el valor medio de la señal de entrada,
la multiplica por 1,11 y muestra el resultado. El resultado es exacto si la señal de entrada es una
onda sinusoidal pura. El modelo DM-100 de Greenlee es un multímetro que responde al valor medio.
El método de lectura de valores eficaces reales utiliza un circuito interno para leer el valor eficaz
real. Este método es exacto, dentro de las limitaciones de factor de cresta especificadas, indepen-
dientemente del tipo de señal de entrada, ya sea una onda sinusoidal pura, rectangular, triangular,
media onda o señal con armónicas. La capacidad para leer valores eficaces reales brinda una mayor
versatilidad de medición. La unidad modelo DM-110 de Greenlee es un medidor de valores eficaces
reales.
La tabla de Formas de onda y Factores de cresta muestra algunas de las señales de CA y sus valores
eficaces reales más comunes.
Formas de onda y Factores de cresta
* El factor de cresta es el cociente de un valor máximo en relación con el valor eficaz;
está representado por la letra griega ξ.
Forma de onda
Valor eficaz real 100 100 100 100
Valor medio 90 100 87 64
*Factor de cresta 1,414 1 1,73 2
(ξ)
32
Botón Oprima momentáneamente para: Notas:
Azul seleccionar CA o CC para
(sin etiqueta) medir corriente o tensión
seleccionar funciones secundarias
RANGE cambiar de selección automática de
escala al modo de selección manual
de escala
seleccionar escalas al utilizar el
modo de selección manual de escala
seleccionar niveles de activación
al medir frecuencia
PWR RST
volver a encender luego de
(DM-100
apagado automático
únicamente)
Hz ver la frecuencia mientras
(DM-110 mide tensión de CA o corriente
únicamente)
HOLD retener un valor medido para todas
las funciones. Oprima nuevamente
para regresar al modo normal
Cómo utilizar las distintas funciones
Oprima y mantenga oprimido
para cambiar de selección
manual de escala a selección
automática de escala.
Se efectuan las conversiones pero
la pantalla no las actualiza.
Mientras se encuentre en el modo
Hz, el botón RANGE no cambia la
escala de frecuencia. Sin embargo,
cambia la escala de tensión.
DM-100 • DM-110
33
Operación
1. Consulte la Tabla de valores. Coloque el interruptor de selección en el valor apropiado, oprima
el botón azul sin etiqueta para seleccionar la medición deseada y conecte los cables de prueba
al medidor. Empiece por el ajuste de medición más alto.
2. Consulte la sección Mediciones más comunes en relación con las instrucciones específicas
para cada tipo de medición.
3. Pruebe la unidad en un circuito o componente que se sabe está funcionando perfectamente.
Si no funciona como debería en un circuito que se sabe está funcionando perfectamente,
reemplace la pila y/o el fusible.
Si sigue sin funcionar como debería, devuélvala a Greenlee a fin de que sea reparada.
Vea la dirección bajo Garantía.
4. Anote la lectura del circuito o componente que se está verificando. Si la resolución de la lectura
no es satisfactoria, desconecte el medidor del circuito y cambie a la siguiente escala inferior si
está utilizando la selección manual de escala.
34
Tabla de valores
conecte y conecte
este icono el cable el cable
Para medir esto... coloque en aparecerá en de prueba de prueba
este símbolo... la pantalla rojo a negro a
Capacitancia* F Hz COM
(DM-110 únicamente) VΩµA
Continuidad** y oprima el botón AZUL VΩµA COM
una sola vez
µA
Corriente (CA) y oprima el botón AZUL µAVΩµA COM
una sola vez
Corriente (CC) µA VΩµA COM
µA
Corriente (CA) 10A
(10 A máx) y oprima el botón AZUL 10 A COM
una sola vez A
Corriente (CC) 10A 10 A COM
(10 A máx) A
Diodo*** y oprima el botón AZUL VVΩµA COM
dos veces
Frecuencia Hz kHz o MHz Hz COM
(DM-110 únicamente) VΩµA
Resistencia M o k VΩµA COM
V
Tensión (CA) y oprima el botón AZUL mV o V VΩµA COM
una sola vez
Tensión (CC) V VΩµA COM
mV o V
* Descargue el condensador antes de efectuar una medición.
** El tono suena si la resistencia medida es menor a 30 aproximadamente.
*** Se indica la tensión de diodo si se encuentra en estado de conducción,
o se muestra OL si se encuentra en estado aislante.
3200
µA máx para DM-100, 4000 µA máx para DM-110.
Operación (continuación)
DM-100 • DM-110
35
Operación (continuación)
Apagado automático
El medidor conmutará por defecto para apagarse automáticamente luego de un período de
inactividad por parte del usuario. en la pantalla indica que la función se ha activado. El
medidor emite un tono audible por espacio de 15 segundos antes de apagarse. El modelo DM-100
se apaga luego de 10 minutos y el modelo DM-110 se apaga luego de 30 minutos. La lectura actual
se retendrá en la memoria. Para volver a encender el medidor, oprima cualquier botón o gire el
interruptor de selección. La última lectura se mostrará en el modo Hold (Retención de datos en
pantalla). Oprima HOLD para regresar a la operación normal.
Desactivación de la función de Apagado automático
Para desactivar la función de Apagado automático, oprima y mantenga oprimido el botón RANGE
al tiempo que enciende la unidad.
Nota: Cuando el medidor se apaga y se vuelve a encender con el interruptor de selección, la unidad
regresará a sus valores predeterminados. Sin embargo, la memoria no retendrá el último valor.
36
Mediciones más comunes
Medición de tensión
Medición de corriente
DM-100 DM-110
37
Mediciones más comunes (continuación)
Medición de resistencia
Verificación de continuidad
38
Mediciones más comunes (continuación)
Medición de capacitancia
Verificación de diodo
Polarización directa
Polarización inversa
DM-100 DM-110
39
Precisión: DM-100
Consulte la sección Especificaciones en relación con las condiciones de operación y el coeficiente
de temperatura.
La precisión se especifica de la siguiente manera: ± (un porcentaje de la lectura + una cantidad fija)
a 23°C a ± 5°C (73,4°F ± 9°F), 0% a 80% de humedad relativa.
Para el modelo DM-100, la precisión de CA se especifica para ondas sinusoidales únicamente.
Corriente alterna (CA)
Escala* Precisión Tensión de carga
320,0 µA ± (2,2% + 0,5 µA) 1V máx
3200 µA ± (2,2% + 5 µA) 10V máx
10,00 A ± (2,5% + 0,05 A) 2V máx
* En el modo de selección automática de escala, la escala cambia a aproximadamente 3.200 puntos.
En el modo de selección manual de escala, la escala se extiende a 3.400 puntos.
Tensión alterna (CA)
Escala* Precisión Escala de frecuencia Impedancia de entrada
3,200V ± (1,50% + 0,005V)
32,00V ± (1,50% + 0,05V)
320,0V ± (1,50% + 0,5V)
750V ± (1,50% + 5V) inal
40 a 500 Hz 10 M, 100 pF
Corriente continua (CC)
Escala* Precisión Tensión de carga
320,0 µA ± (1,7% + 0,2 µA) 1V máx
3200 µA ± (1,7% + 2 µA) 10V máx
10,00 A ± (2% + 0,02 A) 2V máx
40
Precisión: DM-100 (continuación)
Resistencia
Escala* Precisión
320,0 Ω± (1% + 0,4 )
3,200 kΩ± (0,75% + 0,003 k)
32,00 kΩ± (0,75% + 0,03 k)
320,0 kΩ± (0,75% + 0,3 k)
3,200 MΩ± (1% + 0,003 M)
32,00 MΩ± (2% + 0,05 M)
Verificación de diodo
Corriente de prueba (Típica): 1,5 mA
Tensión de circuito abierto: Menor de 3,1V CC
Continuidad
Umbral: El tono suena si la resistencia medida es menor a 30 aproximadamente.
Tiempo de respuesta: Aproximadamente 100 ms.
Tensión de circuito abierto típica: 1,5V
* En el modo de selección automática de escala, la escala cambia a aproximadamente 3.200 puntos.
En el modo de selección manual de escala, la escala se extiende a 3.400 puntos.
Tensión continua (CC)
Escala* Precisión Impedancia de entrada
320,0 mV ± (0,5% + 0,2 mV) superior a 1000 M
3,200V ± (0,5% + 0,002V)
32,00V ± (0,5% + 0,02V)
320,0V ± (0,5% + 0,2V)
1000V ± (0,5% + 2V)
10 M
DM-100 DM-110
41
Precisión: DM-110
Consulte la sección Especificaciones en relación con las condiciones de operación y el coeficiente
de temperatura.
La precisión se especifica de la siguiente manera: ± (un porcentaje de la lectura + una cantidad fija)
a 23°C a ± 5°C (73,4°F ± 9°F), 0% a 80% de humedad relativa.
Para el modelo DM-110, la especificación sobre precisión de CA aplica a ondas sinusoidales a plena
escala y ondas no sinusoidales por debajo de la escala media.
Para factores de cresta de 1,4 a 3, agregue 1,5% a la precisión.
Para factores de cresta de 3 a 4, agregue 3% a la precisión.
Corriente continua (CC)
Escala Precisión Tensión de carga
400,0 µA ± (1,7% + 0,2 µA) 1V máx
4000 µA ± (1,7% + 2 µA) 10V máx
10,00 A ± (2% + 0,02 A) 2V máx
Tensión alterna (CA)
Escala Precisión Escala de frecuencia Impedancia de entrada
400,0 mV no especificado
4,000V ± (1,5% + 0,005V)
40,00V ± (1,3% + 0,05V) 40 a 500 Hz 10 M, 100 pF
400,0V ± (1,3% + 0,5V)
750V ± (1,3% + 5V)
Corriente alterna (CA)
Escala Precisión Tensión de carga
400,0 µA ± (2,2% + 0,5 µA) 1V máx
4000 µA ± (2,2% + 5 µA) 10V máx
10,00 A* ± (2,5% + 0,05 A) 2V máx
* 10 A continuos; 20 A máximo (Régimen de trabajo: 30 segundos encendido,
5 minutos apagado)
42
Precisión: DM-110 (continuación)
Estas precisiones son para los condensadores de película
(condensadores con absorción dieléctrica inapreciable)
*Precisión especificada para valores menores de media escala completa.
Puede haber fluctuacion dentro de los límites de precisión específicos.
Tensión de circuito abierto típica: 1,3V
Resistencia
Escala Precisión
400,0 Ω± (1,0% + 0,3 )
4,000 kΩ± (0,75% + 0,002 k)
40,00 kΩ± (0,75% + 0,02 k)
400,0 kΩ± (0,75% + 0,2 k)
4,000 MΩ± (1,0% + 0,003 M)
40,00 MΩ± (1,5% + 0,05 M)
Capacitancia
Escala Precisión
4,000 nF ± (3,0% + 0,010 nF)
40,00 nF ± (3,0% + 0,10 nF)
400,0 nF ± (2,0% + 0,8 nF)
4,000 µF ± (2,0% + 0,008 µF)
40,00 µF ± (2,0% + 0,08 µF)
400,0 µF ± (2,0% + 0,8 µF)
4,000 mF* ± (5,0% + 0,020 mF)
40,00 mF* ± (5,0% + 0,20 mF)
Tensión continua (CC)
Escala Precisión Impedancia de entrada
400,0 mV ± (0,5% + 0,2 mV) superior a 1000 M
4,000V ± (0,5% + 0,002V)
40,00V ± (0,5% + 0,02V)
400,0V ± (0,5% + 0,2V)
1000V ± (0,5% + 2V)
10 M
DM-100 DM-110
43
Precisión: DM-110 (continuación)
Verificación de diodo
Corriente de prueba (Típica): 1,5 mA
Tensión de circuito abierto: Menor de 3,1V CC
Continuidad
Umbral: El tono suena si la resistencia medida es menor a 30 aproximadamente.
Tiempo de respuesta: Aproximadamente 100 ms.
*La sensitividad es 1,5V rms por debajo de 20 Hz
Con la función de medición de frecuencia se puede seleccionar entre cuatro niveles de activación.
Utilice el botón Range para seleccionar el deseado.
Frecuencia
Escala Precisión Sensitividad
4,000 kHz ± (0,01% + 0,001 kHz)
40,00 kHz ± (0,01% + 0,01 kHz) 100 mV rms *
400,0 kHz ± (0,01% + 0,1 kHz)
4,000 MHz ± (0,01% + 0,001 MHz) 250 mV rms *
40,00 MHz ± (0,01% + 0,01 MHz) 1V rms
44
Especificaciones
Pantalla:
DM-100: cristal líquido (LCD) (3200) y gráfico de barras de 70 segmentos
En el modo de selección automática, la escala cambia a aproximadamente 3200 puntos.
En el modo de selección manual de escala, la escala se extiende a 3400 puntos.
DM-110: cristal líquido (LCD) (4000) y gráfico de barras de 82 segmentos
Polaridad: Automática
Frecuencia de muestreo:
Pantalla numérica: 2 por segundo
Pantalla de gráfico de barras: 12 por segundo
Coeficiente de temperatura: 0,15 x (precisión) por °C menor de 18°C o mayor de 28°C
Apagado automático:
DM-100: Después de 10 minutos de inactividad
DM-110: Después de 30 minutos de inactividad
Protecciones contra sobrecarga:
Voltios: Consulte las Categorías de protección de sobretensión
Amps: A: Fusible16A/500V HBC F, condiciones de corte 10 kA, 1/4 x 1-1/4 pulg.
Otras funciones: 600V CC/V CA eficaces
Categorías de protección de sobretensión: Terminal VΩµA:
Categoría III, 600V CA y 600V CC
Categoría II, 1000V CA y 1000V CC
Condiciones de operación:
0°C a 30°C (32°F a 86°F), 0% a 80% de humedad relativa
30°C a 40°C (86°F a 104°F), 0% a 75% de humedad relativa
40°C a 50°C (104°F a 122°F), 0% a 45% de humedad relativa
Uso en interiores únicamente.
Condiciones de almacenamiento:
-20°C a 60°C (-4°F a 140°F), 0% a 80% de humedad relativa
Retire las pilas
Altura: 2.000 m (6.500 pies) máximo
Grado de contaminación: 2
Pila:
DM-100: Dos pilas AAA de 1,5V (NEDA 24A o IEC LR03)
DM-110: Pila de 9 voltios (NEDA 1604 o IEC 6F22)
Vida útil de la pila:
DM-100: Aproximadamente 500 horas con pilas alcalinas
DM-110: Aproximadamente 300 horas con pilas alcalinas
DM-100 DM-110
45
Categorías de instalación de sobretensión
Las siguientes definiciones proceden de la norma de seguridad internacional sobre la coordinación
de aislamientos tal y como se aplica a equipos de medición, control y laboratorio. En las
publicaciones IEC 1010-1 y IEC 60664 de la International Electrotechnical Commission
(Comisión Electrotécnica Internacional) se detallan más a fondo estas categorías de sobretensión.
Categoría de sobretensión I
Nivel de señal. Equipo electrónico y de telecomunicaciones, o partes del mismo. Como ejemplo
pueden citarse los circuitos electrónicos protegidos contra tensiones momentáneas dentro de
fotocopiadores y modems.
Categoría de sobretensión II
Nivel local. Aparatos eléctricos, equipo portátil, y los circuitos a los que están conectados. Como
ejemplo pueden citarse dispositivos de iluminación, televisores y circuitos de rama larga.
Categoría de sobretensión III
Nivel de distribución. Máquinas instaladas permanentemente y los circuitos a los que están
cableados. Como ejemplo pueden citarse sistemas conductores y los paneles del interruptor
automático principal del sistema eléctrico de un edificio.
Categoría de sobretensión IV
Nivel de abastecimiento primario. Líneas aéreas y otros sistemas de cable. Como ejemplo pueden
citarse cables, medidores, transformadores y cualquier otro equipo exterior perteneciente a la
empresa de servicio eléctrico.
46
Mantenimiento
1. Desconecte la unidad del circuito. Apague (OFF) la unidad.
2. Retire los tornillos de la tapa posterior.
3. Retire la tapa posterior.
4. Levante el compartimiento de la pila de la tapa posterior.
5. Reemplace las pilas (fíjese en la polaridad) y/o el (los) fusible(s).
6. Vuelva a colocar la tapa, los tornillos y el apoyo.
Limpieza
Limpie periódicamente la caja utilizando un paño húmedo y detergente suave; no utilice abrasivos ni
solventes.
No intente reparar esta unidad, ya que contiene partes que deben recibir mantenimiento
por parte de un profesional.
No exponga la unidad a ambientes de temperatura extrema o altos niveles de humedad;
Consulte las Especificaciones.
De no observarse estas precauciones podrían sufrirse lesiones o daños a la unidad.
El fusible es una parte integral para la protección contra sobretensión. Cuando sea necesario
reemplazarlo, consulte la sección Especificaciones para saber qué tipo, tamaño y
capacidad debe tener. Utilizar cualquier otro tipo de fusible anulará la clasificación de
protección de sobretensión de la unidad.
De no observarse esta advertencia pueden sufrirse graves lesiones o incluso la muerte.
Cómo reemplazar la pila y el fusible
Antes de abrir la caja, retire del circuito los cables de prueba y apague la unidad.
De no observarse esta advertencia pueden sufrirse graves lesiones o incluso la muerte.
DM-100 DM-110
47
Greenlee Textron / Subsidiary of Textron Inc. 4455 Boeing Dr., Rockford, IL 61109-2988 815/397-7070
Greenlee Textron / Subsidiary of Textron Inc. 4455 Boeing Dr., Rockford, IL 61109-2988 815/397-7070
DM-100 • DM-110
49
Description
Les multimètres numériques DM-100 et DM-110 sont des appareils de vérification portables dotés
des capacités de mesure suivantes : la tension c.a. et c.c., le courant c.a. et c.c. et la résistance. Ils
contrôlent aussi les diodes et vérifient la continuité. En outre, le DM-110 mesure la fréquence et la
capacité. Le DM-110 est un multimètre à lecture RMS véritable.
Les deux modèles sont équipés d’une fonction de mise hors tension automatique qui économise les
piles et d’un affichage graphique plus sensible que l’affichage numérique — utile pour détecter les
contacts défectueux, les clics de potentiomètre et les brusques variations de signal.
Sécurité
Lors de l’utilisation et de l’entretien des outils et des équipements de Greenlee, votre sécurité est une
priorité. Ce manuel d’instructions et toute étiquette sur l’outil fournit des informations permettant
d’éviter des dangers ou des manipulations dangereuses liées à l’utilisation de cet outil. Suivre
toutes les consignes de sécurité indiquées.
Dessein
Ce manuel d’instructions est conçu pour que le personnel puisse se familiariser avec le
fonctionnement et les procédures d’entretien sûres des multimètres numériques DM-100 et
DM-110 de Greenlee.
Mettre ce manuel à la disposition de tous les employés.
On peut obtenir des exemplaires gratuits sur simple demande.
Greenlee et sont des marques déposées de Greenlee Textron.
CONSERVER CE MANUEL
50
Consignes de sécurité importantes
Lire attentivement et bien comprendre cette documentation avant
d’utiliser ou de procéder à l’entretien de cet équipement. Négliger
de comprendre comment utiliser cet outil en toute sécurité peut
provoquer un accident et entraîner des blessures graves, voire
mortelles.
Ce symbole met en garde contre les risques et les manipulations dangereuses pouvant
entraîner des blessures ou l’endommagement du matériel. Le mot indicateur, défini ci-
dessous, indique la gravité du danger. Le message qui suit le mot indicateur indique
comment empêcher le danger.
SYMBOLE D’AVERTISSEMENT
Danger immédiat qui, s’il n’est pas pris en considération ENTRAINERA des blessures
graves, voire mortelles.
Danger qui, s’il n’est pas pris en considération, POURRAIT entraîner des blessures graves,
voire mortelles.
Dangers ou manipulations dangereuses qui, s’ils ne sont pas pris en considération,
POURRAIENT EVENTUELLEMENT entraîner des dommages à la propriété ou causer des
blessures.
DM-100 • DM-110
51
Consignes de sécurité importantes
Risques de décharge électrique et d’incendie :
Ne pas exposer cet appareil à la pluie ou à l’humidité.
Ne pas utiliser cet appareil s’il est mouillé ou endommagé.
Utiliser des fils d’essai ou des accessoires conformes à l’application. Consulter la
catégorie et la tension nominale du fil d’essai ou de l’accessoire.
Vérifier les fils d’essai ou l’accessoire avant de les utiliser. La pièce (ou les pièces)
doi(ven)t être propre(s) et sèche(s) et l’isolation en bon état.
Utiliser cet appareil uniquement dans le but pour lequel il a été conçu, tel que décrit dans
ce manuel. Toute autre utilisation peut altérer le système de protection de cet appareil.
L’inobservation de ces consignes peut entraîner des blessures graves, voire mortelles.
Ne pas utiliser lorsque le boîtier est ouvert.
Avant d’ouvrir le boîtier, retirer les fils d’essai du circuit et mettre l’appareil hors tension.
L’inobservation de ces consignes peut entraîner des blessures graves, voire mortelles.
Risques de décharge électrique :
Un contact avec des circuits sous tension peut entraîner des blessures
graves, voire mortelles.
Risques de décharge électrique :
Ne pas appliquer plus que la tension nominale entre deux bornes d’entrée, ou entre une
borne d’entrée et une prise de terre.
Ne pas entrer en contact avec les extrémités des fils d’essai ou avec toute autre partie non
isolée de l’accessoire.
L’inobservation de ces consignes peut entraîner des blessures graves, voire mortelles.
52
Consignes de sécurité importantes
Ne pas tenter de réparer cet appareil. Il ne comporte aucune pièce pouvant être réparée.
Ne pas exposer l’appareil à des températures ou à une humidité extrêmes.
Voir les spécifications.
L’inobservation de ces consignes peut endommager l’appareil et entraîner des blessures.
Ne pas modifier la fonction de mesure pendant que les fils d’essai sont connectés à un
composant ou à un circuit.
L’inobservation de cette consigne peut endommager l’appareil et entraîner des blessures.
Le fusible est une partie intégrante du système de protection de surtension. Lorsqu’un
fusible doit être remplacé, consulter les spécifications pour connaître le type, la taille et la
capacité requis. L’utilisation de tout autre type de fusible annule le calibrage de protection
contre la surtension de l’appareil.
L’inobservation de cette consigne peut entraîner des blessures graves, voire mortelles.
Sauf si l’on mesure la tension, le courant ou la fréquence, mettre hors tension et couper
la source d’alimentation. S’assurer que tous les condensateurs sont déchargés. Aucune
tension ne doit être présente.
Régler le sélecteur et connecter les fils d’essai pour qu’ils correspondent à la mesure
voulue. Des réglages ou des connexions incorrects peuvent faire sauter les fusibles.
L’utilisation de cet appareil à proximité d’équipements qui génèrent des interférences
électromagnétiques peut produire des lectures instables ou erronées.
DM-100 • DM-110
53
Identification
1. Affichage Graphique à barres sur cristaux liquides avec 4 chiffres
2. Boutons de fonction Voir les explications de la section Utilisation des fonctions
3. Sélecteur Sélectionne une fonction ou met hors tension (OFF).
4. Hz Borne d’entrée positive pour toutes les mesures
sauf
pour le courant
VΩµA de 10 A ( Hz sur le DM-110 uniquement)
5. COM Borne d’entrée négative, commune ou reliée à la terre, pour toutes les
mesures
6. 10A Borne d’entrée positive pour les mesures de courant allant jusqu’à 10 A
1
2
5
6
3
4
Symboles apparaissant sur lappareil
Avertissement — Lire le manuel d’instructions
Pile
Fusible
Isolation double
54
Icônes daffichage
Les deux modèles
7. Plage manuelle est sélectionnée
8. Mesure du c.c. sélectionnée
9. Mesure du c.a. sélectionnée
10. Pile faible ou déchargée
11. APO
Mise hors tension automatique activée
12. Diode
13. Continuité est sélectionné
14. µ Micro (10
-6
)
15. m Milli (10
-3
)
16. V Volts
17. A Ampères
18. M Méga (10
6
)
19. k Kilo (10
3
)
20. Ohms
21. Hz Hertz (fréquence ou
cycles par seconde)
7
9
8
10
22
14,15,16,17
18,19,20,21
12 1311
7
8910
1123 12
13
14,15,16,17
24,25,26,27
18,19,20,21
DM-100 uniquement
22.
Fonction de maintien activée
DM-110 uniquement
23.
Fonction de maintien activée
24. n nano (10
-9
)
25. µ Micro (10
-6
)
26. m Milli (10
-3
)
27. F Farads
Remarque : Les icônes non identifiées
ne sont pas utilisées
sur ces modèles.
Affichage du DM-100
Affichage du DM-110
DM-100 • DM-110
55
Mesure du c.a.
En règle générale, les mesures du c.a. sont affichées en valeurs RMS (moyenne quadratique). Les
deux méthodes de mesure du c.a. sont calibrées pour une réponse RMS moyenne et pour une
lecture RMS véritable.
La méthode par valeur moyenne consiste à mesurer la valeur moyenne du signal d’entrée, à la
multiplier par 1,11 et à afficher le résultat. Cette méthode est précise si le signal d’entrée est une
onde sinusoïdale pure. Le Greenlee DM-100 mesure la réponse moyenne.
La méthode par mesure de la valeur efficace vraie (RMS) fait appel à un circuit interne permettant
de mesurer directement la valeur efficace. Cette méthode est précise dans les limites du facteur
de crête spécifiées, que le signal soit une onde pure, une onde carrée, une onde triangulaire, une
demi-onde ou un signal comportant des harmoniques. Les appareils à mesure de valeur efficace
vraie (RMS) sont beaucoup plus polyvalents que les appareils conventionnels. Le multimètre
DM-110 de Greenlee lit la valeur RMS véritable.
Le tableau des formes d’onde et facteurs de crête fournit les valeurs efficaces des signaux alternatifs
courants.
Formes dondes et facteurs de crêtes
* Le facteur de crête correspond au ratio de la valeur de crête par rapport à la valeur RMS ; il est
représenté par la lettre grecque ξ.
Forme donde
Valeur RMS 100 100 100 100
Valeur moyenne 90 100 87 64
Facteur de crête* 1,414 1 1,73 2
(ξ)
56
Bouton Appuyer quelques secondes pour : Remarques :
Bleu sélectionner c.a. ou c.c. lors de la
(pas d’étiquette) mesure du courant ou de la tension
sélectionner les fonctions secondaires
RANGE changer à partir de la mesure
automatique de la plage à plage
manuelle
sélectionner les plages lors de
l’utilisation du mode de plage
manuelle
sélectionner les niveaux de
déclenchement pendant la fréquence
de mesure
PWR RST
remettre en marche après l’extinction
(DM-100
auto
uniquement)
Hz visionner la fréquence pendant
(DM-110 la mesure de la tension c.a. ou
uniquement)
du courant
HOLD maintenir une valeur mesurée pour
toutes les fonctions. Presser de
nouveau pour revenir au mode normal
Utilisation des fonctions
Les conversions sont effectuées
mais l’affichage ne se met pas
à jour.
Presser et maintenir pour
changer la sélection de plage
de manuelle à automatique.
En mode Hz, le bouton RANGE
ne change pas la plage de
fréquence. Toutefois, il change
la plage de tension.
DM-100 • DM-110
57
Utilisation
1. Se reporter au tableau des réglages. Régler le sélecteur au réglage approprié, appuyer sur le
bouton bleu sans étiquette pour sélectionner la mesure voulue et connecter les fils d’essai au
multimètre. Commencer avec le réglage de mesures le plus élevé.
2. Se reporter aux mesures types pour obtenir des instructions de mesure spécifiques.
3. Vérifier l’appareil sur un circuit ou sur un composant connu.
Si l’appareil ne fonctionne pas comme prévu sur un circuit dont le fonctionnement est
connu, remplacer la pile et/ou le fusible.
Si l’appareil ne fonctionne toujours pas comme prévu, le renvoyer à Greenlee pour qu’il
soit réparé. Voir l’adresse indiquée sous la garantie.
4. Lire le circuit ou le composant à vérifier. Si la résolution n’est pas satisfaisante, retirer le
multimètre du circuit et passer à la plage immédiatement moins élevée à l’aide de la sélection
manuelle de la plage.
58
Tableau des réglages
régler le sélecteur
cette icône connecter et connecter
Pour mesurer cela ...
à ce symbole
apparaît sur le fil le fil
sur l’écran rouge à… noir à ...
Capacité* F Hz COM
(DM-110 uniquement) VΩµA
Continuité** et appuyer sur le bouton VΩµA COM
BLEU une fois
µA
Courant (c.a.)† et appuyer sur le bouton µAVΩµA COM
BLEU une fois
Courant (c.c.)† µA VΩµA COM
µA
Courant (c.a.) 10A
(10 A max.) et appuyer sur le bouton 10 A COM
BLEU une fois A
Courant (c.c.) 10A 10 A COM
(10 A max.) A
Diode*** et appuyer sur le bouton VVΩµA COM
BLEU deux fois
Fréquence Hz kHz ou MHz Hz COM
(DM-110 uniquement) VΩµA
Résistance M ou k VΩµA COM
V
Tension (c.a.) et appuyer sur le bouton mV ou V VΩµA COM
BLEU une fois
Tension (c.c.) V VΩµA COM
mV ou V
* Décharger le condensateur avant de mesurer.
** Une tonalité se fait entendre si la résistance mesurée est inférieure à environ 30 .
*** La tension de la diode est illustrée s’il y conductance, ou « OL » est affiché s’il n’y a pas
conductance.
3 200
µA max. pour le DM-100, 4 000 µA max. pour le DM-110.
Utilisation (suite)
DM-100 • DM-110
59
Utilisation (suite)
Mise hors tension automatique
Le multimètre s’éteint automatiquement après une certaine période d’inactivité. sur
l’affichage indique que cette fonction est activée. Le multimètre émet un bip environ 15 secondes
avant de s’éteindre. Le DM-100 s’éteint après environ 10 minutes et le DM-110 après 30 minutes.
La lecture en cours est gardée en mémoire. Pour rallumer le multimètre, appuyer sur n’importe
quel bouton ou tourner le sélecteur. La dernière lecture s’affiche en mode de maintien. Appuyer
sur HOLD pour revenir au mode normal de fonctionnement.
Mise hors tension automatique
Pour désactiver la fonction de mise hors tension automatique, appuyer sur et maintenir le bouton
RANGE en allumant le multimètre.
Remarque : Lorsque le multimètre est mis hors tension puis rallumé de nouveau à l’aide du sélecteur,
l’appareil revient au mode par défaut. Toutefois, la dernière valeur n’est pas gardée en mémoire.
60
Mesures types
Mesure de la tension
Mesure du courant
DM-100 DM-110
61
Mesures types (suite)
Mesure de la résistance
Vérification de la continuité
62
Mesures types (suite)
Mesure de la capacité
Mesure des diodes
Polarisation Directe
Polarité Inversée
DM-100 DM-110
63
Précision : DM-100
Pour obtenir les conditions d’utilisation et les coefficients de température, consulter la section sur
les spécifications.
La précision est spécifiée comme suit : ± (un pourcentage de la lecture + une quantité fixe) à 23 °C
± 5 °C (73,4 °F ± 9 ºF), 0 à 80 % d’humidité relative.
Pour le DM-100, la précision du c.a. est spécifiée uniquement pour les ondes sinusoïdales.
Courant c.a.
Plage* Précision Tension de consommation
320,0 µA ± (2,2 % + 0,5 µA) 1 V max
3 200 µA ± (2,2 % + 5 µA) 10 V max
10,00 A ± (2,5 % + 0,05 A) 2 V max
* En mode de sélection automatique de plage, les plages changent à environ tous les 3 200 comptes.
En mode de sélection de plage manuel, la plage s’étend jusqu’à 3 400 comptes.
Tension c.a.
Plage* Précision Plage de fréquence Impédance dentrée
3,200 V ± (1,50 % + 0,005 V)
32,00 V ± (1,50 % + 0,05 V)
320,0 V ± (1,50 % + 0,5 V)
750 V ± (1,50 % + 5 V)
40 à 500 Hz 10 M, 100 pF
Courant c.c.
Plage* Précision Tension de consommation
320,0 µA ± (1,7 % + 0,2 µA) 1 V max
3 200 µA ± (1,7 % + 2 µA) 10 V max
10,00 A ± (2 % + 0,02 A) 2 V max
64
Précision : DM-100 (suite)
Résistance
Plage* Précision
320,0 Ω± (1 % + 0,4 )
3,200 kΩ± (0,75 % + 0,003 k)
32,00 kΩ± (0,75 % + 0,03 k)
320,0 kΩ± (0,75 % + 0,3 k)
3,200 MΩ± (1 % + 0,003 M)
32,00 MΩ± (2 % + 0,05 M)
Vérification de la diode
Courant de vérification (Typique) : 1,5 mA
Tension de circuit ouvert : Moins de 3,1 V c.c.
Continuité
Seuil : Une tonalité se fait entendre si la résistance mesurée est inférieure à environ 30 .
Temps de réponse : Environ 100 ms
Tension typique de circuit ouvert : 1,5 V
* En mode de sélection automatique de plage, les plages changent à environ tous les 3 200 comptes.
En mode de sélection de plage manuel, la plage s’étend jusqu’à 3 400 comptes.
Tension c.c.
Plage* Précision Impédance dentrée
320,0 mV ± (0,5 % + 0,2 mV) supérieur à 1 000 M
3,200 V ± (0,5 % + 0,002 V)
32,00 V ± (0,5 % + 0,02 V)
320,0 V ± (0,5 % + 0,2 V)
1 000 V ± (0,5 % + 2 V)
10 M
DM-100 DM-110
65
Précision : DM-110
Pour obtenir les conditions d’utilisation et les coefficients de température, consulter la section sur
les spécifications.
La précision est spécifiée comme suit : ± (un pourcentage de la lecture + une quantité fixe) à 23 °C
± 5 °C (73,4 °F ± 9 ºF), 0 à 80 % d’humidité relative.
Pour le DM-110, la spécification de précision du c.a. s’applique aux ondes sinusoïdales à leur pleine
échelle et aux ondes non-sinusoïdales inférieures à la demi-grandeur.
Pour les facteurs de crête entre 1,4 et 3, ajouter 1,5 % à la précision.
Pour les facteurs de crête entre 3 et 4, ajouter 3 % à la précision.
Courant c.c.
Plage Précision Tension de consommation
400,0 µA ± (1,7 % + 0,2 µA) 1 V max
4 000 µA ± (1,7 % + 2 µA) 10 V max
10,00 A ± (2 % + 0,02 A) 2 V max
Tension c.a.
Plage Précision Plage de fréquence Impédance dentrée
400,0 mV non spécifié
4 000 V ± (1,5 % + 0,005 V)
40,00 V ± (1,3 % + 0,05 V) 40 à 500 Hz 10 M, 100 pF
400,0 V ± (1,3 % + 0,5 V)
750 V ± (1,3 % + 5 V)
Courant c.a.
Plage Précision Tension de consommation
400,0 µA ± (2,2 % + 0,5 µA) 1 V max
4 000 µA ± (2,2 % + 5 µA) 10 V max
10,00 A* ± (2,5 % + 0,05 A) 2 V max
* 10 A continu ; 20 A maximum (Durée de mise sous tension : 30 secondes sous
tension, 5 minutes hors tension)
66
Précision : DM-110 (suite)
Les précisions valent pour les condensateurs à couches (condensateurs dont l’absorption
diélectrique est négligeable).
*La précision spécifiée pour les valeurs inférieures à la moitié de la pleine échelle.
Plusieurs peuvent fluctuer aux environs des limites de précision spécifiées.
Tension typique de circuit ouvert : 1,3 V
Résistance
Plage Précision
400,0 Ω± (1,0 % + 0,3 )
4,000 kΩ± (0,75 % + 0,002 k)
40,00 kΩ± (0,75 % + 0,02 k)
400,0 kΩ± (0,75 % + 0,2 k)
4,000 MΩ± (1,0 % + 0,003 M)
40,00 MΩ± (1,5 % + 0,05 M)
Capacité
Plage Précision
4,000 nF ± (3,0 % + 0,010 nF)
40,00 nF ± (3,0 % + 0,10 nF)
400,0 nF ± (2,0 % + 0,8 nF)
4,000 µF ± (2,0 % + 0,008 µF)
40,00 µF ± (2,0 % + 0,08 µF)
400,0 µF ± (2,0 % + 0,8 µF)
4,000 mF* ± (5,0 % + 0,020 mF)
40,00 mF* ± (5,0 % + 0,20 mF)
Tension c.c.
Plage Précision Impédance dentrée
400,0 mV ± (0,5 % + 0,2 mV) supérieur 1 000 M
4,000 V ± (0,5 % + 0,002 V)
40,00 V ± (0,5 % + 0,02 V)
400,0 V ± (0,5 % + 0,2 V)
1 000 V ± (0,5 % + 2 V)
10 M
DM-100 DM-110
67
Précision : DM-110 (suite)
Vérification de la diode
Courant de vérification (Typique) : 1,5 mA
Tension de circuit ouvert : Moins de 3,1 V c.c.
Continuité
Seuil : Une tonalité se fait entendre si la résistance mesurée est inférieure à environ 30 .
Temps de réponse : Environ 100 ms
*La sensibilité est de 1,5 V rms inférieure à 20 Hz
La fonction de mesure de la fréquence comporte quatre sélections du niveau de la gâchette.
Utiliser le bouton « Range » pour sélectionner.
Fréquence
Plage Précision Sensibilité
4,000 kHz ± (0,01 % + 0,001 kHz)
40,00 kHz ± (0,01 % + 0,01 kHz) 100 mV rms *
400,0 kHz ± (0,01 % + 0,1 kHz)
4,000 MHz ± (0,01 % + 0,001 MHz) 250 mV rms
40,00 MHz ± (0,01 % + 0,01 MHz) 1 V rms
68
Spécifications
Affichage :
DM-100 : Affichage à cristaux liquides (3 200) et graphique à barres à 70 segments
En mode de sélection automatique de plage, les plages changent à environ tous les 3 200
comptes. En mode de sélection de plage manuel, la plage s’étend jusqu’à 3 400 comptes.
DM-110 : Affichage à cristaux liquides (4 000) et graphique à barres à 82 segments
Polarité : Automatique
Taux d’échantillonnage :
Afficheur numérique : 2 par seconde
Afficheur de graphique à barres : 12 par seconde
Coefficient de température : 0,15 x (précision) par °C au-dessous de 18 °C ou au-dessus de 28 °C
Mise hors tension automatique :
DM-100 : après 10 minutes d’inactivité
DM-110 : après 30 minutes d’inactivité
Protections de surcharge :
Volts : Consulter les Catégories de protection contre la surtension
Ampères : A : Fusible F de 16 A/500 V HBC, pouvoir de coupure nominal 10 kA, 1/4 x 1-1/4 po
Autres fonctions : 600 V c.c./V c.a. RMS
Catégories de protection contre la surtension : Borne VΩµA:
Catégorie III, 600 V c.a. et 600 V c.c.
Catégorie II, 1 000 V c.a. et 1 000 V c.c.
Conditions d’utilisation :
0 à 30 °C (32 à 86 °F), 0 à 80 % d’humidité relative
30 à 40 °C (86 à 104 °F), 0 à 75 % d’humidité relative
40 à 50 °C (104 à 122 °F), 0 à 45 % d’humidité relative
Utilisation à l’intérieur uniquement.
Conditions d’entreposage :
-20 à 60 °C (-4 à 140 °F), 0 à 80 % d’humidité relative
Enlever les piles
Elévation : 2 000 m (6 500 pi) maximum
Degré de pollution : 2
Pile :
DM-100 : deux piles AAA de 1,5 volt (NEDA 24 A ou IEC LR03)
DM-110 : une pile de 9 volts (NEDA 1604 ou IEC 6F22)
Durée de vie de la pile :
DM-100 : Environ 500 heures avec une pile alcaline
DM-110 : Environ 300 heures avec une pile alcaline
DM-100 DM-110
69
Catégories dinstallation de surtension
Ces définitions sont dérivées des normes internationales sur la sécurité pour la coordination
de l’isolation telle qu’elle s’applique à la mesure, au contrôle et à l’équipement de laboratoire.
Ces catégories de surtension sont expliquées plus en détail par Commission électrotechnique
internationale ; se reporter à l’une de ces deux publications IEC 1010-1 ou IEC 60664.
Surtension, catégorie I
Niveau de signal. Pièces ou équipement électronique et de télécommunication. Par exemple,
les circuits électroniques protégés contre les courants transitoires, dans les photocopieurs
et les modems.
Surtension, catégorie II
Niveau local. Appareils, équipement portatif et les circuits dans lesquels ils sont branchés. Par
exemple, les appareils d’éclairage, les téléviseurs et les dérivations.
Surtension, catégorie III
Niveau de distribution. Les machines installées en permanence et les circuits auxquels elles sont
câblées. Par exemple, les systèmes de convoyeurs et les panneaux principaux de disjoncteurs du
système électrique d’un édifice.
Surtension, catégorie IV
Niveau d’alimentation principal. Lignes surélevées et autres systèmes de câbles. Par exemple, les
câbles, les multimètres, les transformateurs et autres équipements extérieurs appartenant aux
fournisseurs en électricité.
70
Entretien
1. Débrancher l’appareil du circuit. Mettre l’appareil hors tension (OFF).
2. Enlever les vis du couvercle arrière.
3. Retirer le couvercle arrière.
4. Lever le support de pile pour le sortir du couvercle arrière.
5. Remplacer les piles (selon la polarité) et/ou le fusible.
6. Replacer le couvercle, les vis et le support.
Nettoyage
Nettoyer régulièrement le boîtier avec un chiffon humide et un détergent doux. Ne pas utiliser de
produits abrasifs ou de solvants.
Ne pas tenter de réparer cet appareil. Il ne comporte aucune pièce pouvant être réparée.
Ne pas exposer l’appareil à des températures ou à une humidité extrêmes.
Voir les spécifications.
L’inobservation de ces consignes peut endommager l’appareil et entraîner des blessures.
Le fusible est une partie intégrante du système de protection de surtension. Lorsqu’un
fusible doit être remplacé, consulter les spécifications pour connaître le type, la taille et la
capacité requis. L’utilisation de tout autre type de fusible annule le calibrage de protection
contre la surtension de l’appareil.
L’inobservation de cette consigne peut entraîner des blessures graves, voire mortelles.
Remplacement de la pile et du fusible
Avant d’ouvrir le boîtier, retirer les fils d’essai du circuit et mettre l’appareil hors tension.
L’inobservation de cette consigne peut entraîner des blessures graves, voire mortelles.
DM-100 DM-110
71
Greenlee Textron / Subsidiary of Textron Inc. 4455 Boeing Dr., Rockford, IL 61109-2988 815/397-7070
For technical assistance: 800/435-0786
Greenlee Textron / Subsidiary of Textron Inc.
4455 Boeing Drive, Rockford, IL 61109-2988 USA
Technical / Customer Service (International): 815/397-7070 • Fax: 815/397-9247
Customer Service (North America): 800/435-0786 • Fax: 800/451-2632, 815/397-1865
Canada Fax: 800/524-2853
Lifetime Limited Warranty
Greenlee warrants to the original purchaser of these goods for use that these products will be free from defects in
workmanship and material for their useful life, excepting normal wear and abuse. This warranty is subject to the same
terms and conditions contained in Greenlee’s standard one-year limited warranty.
For all Test Instrument repairs, ship units Freight Prepaid to:
Greenlee Textron, 4411 Boeing Drive, Rockford, IL 61109-2932 USA.
Mark all packages: Attention TEST INSTRUMENT REPAIR. For items not covered under warranty (such as dropped,
abused, etc.), repair cost quote available upon request.
Note: Prior to returning any test instrument, please check replaceable batteries or make sure the battery is at full charge.
Garantía limitada de por vida
Greenlee garantiza al comprador original de estos productos para su uso que estos productos estarán libres de
defectos de mano de obra y materiales durante toda su vida útil, exceptuando el desgaste normal y el abuso. Esta
garantía está sujeta a los mismos términos y condiciones contenidos en la garantía estándar limitada de Greenlee de
un año de duración.
Para reparación de instrumentos de verificación, envíe las unidades con flete pagado a:
Greenlee Textron, 4411 Boeing Drive, Rockford, IL 61109-2932 EE. UU.
Marque todos los paquetes: Atención TEST INSTRUMENT REPAIR (Reparación de instrumentos de verificación). Para
artículos no cubiertos por la garantía (tales como los que se han dejado caer o han sido maltratados, etc.) se puede
cotizar el costo de la reparación a pedido.
Nota: Antes de enviar cualquier instrumento de verificación, revise por favor las pilas o asegúrese de que estén
totalmente cargadas.
Garantie à vie limitée
La société Greenlee garantit à l’acheteur d’origine de ces produits que ces derniers ne comportent aucun défaut
d’exécution ou de matériau pour la durée de leur vie utile, sauf l’usure normale. Cette garantie est assujettie aux mêmes
conditions que celles contenues dans les modalités et conditions de la garantie limitée standard d’un an de Greenlee.
Pour toutes les réparations d’instruments de mesure, expédiez l’appareil en port payé à l’adresse suivante :
Greenlee Textron, 4411 Boeing Drive, Rockford, IL 61109-2932, États-Unis.
Sur tous les colis, inscrivez : Attention : TEST INSTRUMENT REPAIR (Réparation d'instrument de mesure). Lorsque
les articles ne sont pas protégés par une garantie (comme si l’appareil est échappé, s’il est soumis à un usage abusif,
etc.), une soumission pour le prix de réparation sera présentée sur demande.
Remarque : Avant de renvoyer un appareil de mesure, vérifiez la pile et assurez-vous qu’elle est chargée au complet.

Transcripción de documentos

INSTRUCTION INSTRUCTION MANUAL MANUAL MANUAL DE INSTRUCCIONES MANUEL D’INSTRUCTIONS DM-100 • DM-110 DIGITAL MULTIMETERS MULTÍMETROS DIGITALES MULTIMETRES NUMERIQUES Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Lea y entienda todas las instrucciones y la información sobre seguridad que aparecen en este manual, antes de manejar estas herramientas o darles mantenimiento. Lire attentivement et bien comprendre toutes les instructions et les informations sur la sécurité de ce manuel avant d‘utiliser ou de procéder à l‘entretien de cet outil. 999 3788.3 © 2000 Greenlee Textron IM 1463 REV 1 12/00 Description The Greenlee DM-100 and DM-110 Digital Multimeters are hand-held testing devices with the following measurement capabilities: AC and DC voltage, AC and DC current, and resistance. They also check diodes and verify continuity. In addition, the DM-110 measures frequency and capacitance. The DM-110 is a true RMS reading meter. Both models feature automatic power off to conserve the batteries and a bar graph display which is more sensitive than the numeric display—useful for detecting faulty contacts, potentiometer clicks, and signal spikes. Safety Safety is essential in the use and maintenance of Greenlee tools and equipment. This instruction manual and any markings on the tool provide information for avoiding hazards and unsafe practices related to the use of this tool. Observe all of the safety information provided. Purpose This instruction manual is intended to familiarize all personnel with the safe operation and maintenance procedures for Greenlee DM-100 and DM-110 Digital Multimeters. Keep this manual available to all personnel. Replacement manuals are available upon request at no charge. Greenlee and are registered trademarks of Greenlee Textron. KEEP THIS MANUAL 2 DM-100 • DM-110 Important Safety Information SAFETY ALERT SYMBOL This symbol is used to call your attention to hazards or unsafe practices which could result in an injury or property damage. The signal word, defined below, indicates the severity of the hazard. The message after the signal word provides information for preventing or avoiding the hazard. Immediate hazards which, if not avoided, WILL result in severe injury or death. Hazards which, if not avoided, COULD result in severe injury or death. Hazards or unsafe practices which, if not avoided, MAY result in injury or property damage. Read and understand this material before operating or servicing this equipment. Failure to understand how to safely operate this tool can result in an accident causing serious injury or death. 3 Important Safety Information Electric shock hazard: Contact with live circuits can result in severe injury or death. Electric shock and fire hazard: • Do not expose this unit to rain or moisture. • Do not use the unit if it is wet or damaged. • Use test leads or accessories that are appropriate for the application. See the category and voltage rating of the test lead or accessory. • Inspect the test leads or accessory before use. They must be clean and dry, and the insulation must be in good condition. • Use this unit for the manufacturer’s intended purpose only, as described in this manual. Any other use can impair the protection provided by the unit. Failure to observe these warnings can result in severe injury or death. Electric shock hazard: • Do not apply more than the rated voltage between any two input terminals, or between any input terminal and earth ground. • Do not contact the test lead tips or any uninsulated portion of the accessory. Failure to observe these warnings can result in severe injury or death. • Do not operate with the case open. • Before opening the case, remove the test leads from the circuit and shut off the unit. Failure to observe these warnings can result in severe injury or death. 4 DM-100 • DM-110 Important Safety Information The fuse is an integral part of the overvoltage protection. When fuse replacement is necessary, see Specifications for the correct type, size, and capacity. Using any other type of fuse will void the overvoltage protection rating of the unit. Failure to observe this warning can result in severe injury or death. Do not change the measurement function while the test leads are connected to a component or circuit. Failure to observe this precaution can result in injury and can damage the unit. • Do not attempt to repair this unit. It contains no user-serviceable parts. • Do not expose the unit to extremes in temperature or high humidity. See Specifications. Failure to observe these precautions can result in injury and can damage the unit. • Unless measuring voltage, current, or frequency, shut off and lock out power. Make sure that all capacitors are discharged. Voltage must not be present. • Set the selector and connect the test leads so that they correspond to the intended measurement. Incorrect settings or connections can result in a blown fuse. • Using this unit near equipment that generates electromagnetic interference can result in unstable or inaccurate readings. 5 Identification 1. Display 4-digit LCD and bar graph 2. Function Buttons See explanations under Using the Features 3. Selector Selects a function or turns power OFF 4. Hz VΩµA Positive input terminal for all measurements except 10 A current ( Hz on DM-110 only) 5. COM Negative, common or ground input terminal for all measurements 6. 10A Positive input terminal for current measurements up to 10 A 1 2 3 5 4 6 Symbols on the Unit Warning—Read the instruction manual Battery Fuse Double insulation 6 DM-100 • DM-110 Display Icons Both Models DM-100 Only 7. Manual range is selected 8. DC measurement is selected 9. AC measurement is selected 10. Low battery 11. APO Auto Power Off is enabled 12. Diode 13. Continuity is selected 14. µ Micro (10-6) 15. m Milli (10-3) 16. V Volts 17. A Amps 18. M Mega (106) 19. k Kilo (103) 20. Ω Ohms 21. Hz Hertz (frequency or cycles per second) 22. Hold function is enabled DM-110 Only 23. Hold function is enabled 24. n nano (10-9) 25. µ Micro (10-6) 26. m Milli (10-3) 27. F Farads Note: Unidentified icons are not used on these models. DM-110 display 7 23 11 12 13 14,15,16,17 24,25,26,27 18,19,20,21 10 8 9 12 13 DM-100 display 22 11 7 8 9 10 14,15,16,17 18,19,20,21 7 AC Measurement AC measurements are usually displayed as RMS (root-mean-square) values. Two AC measurement methods are average-responding RMS calibrated and true RMS-reading. The average-responding RMS calibrated method takes the average value of the input signal, multiplies it by 1.11, and displays the result. This method is accurate if the input signal is a pure sine wave. The Greenlee DM-100 is an average-responding meter. The true RMS-reading method uses internal circuitry to read the true RMS value. This method is accurate, within the specified crest factor limitations, whether the input signal is a pure sine wave, a square wave, triangle wave, half wave or signal with harmonics. The ability to read true RMS provides much more measurement versatility. The Greenlee DM-110 is a true RMS meter. The Waveforms and Crest Factors table shows some typical AC signals and their RMS values. Waveforms and Crest Factors Waveform RMS Value Average Value Crest Factor* (ξ) 100 90 100 100 100 87 100 64 1.414 1 1.73 2 * The crest factor is the ratio of the peak value to the RMS value; it is represented by the Greek letter ξ. 8 DM-100 • DM-110 Using the Features Button Blue (no label) Press momentarily to: Notes: • select AC or DC when measuring current or voltage • select secondary functions RANGE • change from autoranging to manual ranging • select ranges when using the manual ranging mode • select trigger levels when measuring frequency Press and hold to change from manual ranging to autoranging. PWR RST (DM-100 only) turn power back on after auto power off Hz (DM-110 only) view the frequency while measuring AC voltage or current While in the Hz mode, the RANGE button does not change the frequency range. However, it changes the voltage range. hold a measured value for all functions. Press again to return to normal mode Conversions are made but the display does not update. HOLD 9 Operation 1. See the Settings Table. Set the selector to the proper setting, press the unlabeled blue button to select the desired measurement, and connect the test leads to the meter. Start with the highest measurement setting. 2. See Typical Measurements for specific measurement instructions. 3. Test the unit on a known functioning circuit or component. • If the unit does not function as expected on a known functioning circuit, replace the battery and/or fuse. • If the unit still does not function as expected, send the unit to Greenlee for repair. See the address shown under Warranty. 4. 10 Take the reading from the circuit or component to be tested. If the resolution is not satisfactory, remove the meter from the circuit and change to the next lower range if using manual ranging. DM-100 • DM-110 Operation (cont’d) Settings Table To measure this … set the selector to this symbol … Capacitance* (DM-110 only) F Continuity** Ω and press the BLUE button one time Current (AC)† µA and press the BLUE button one time µA Current (DC)† Current (AC) (10 A max) this icon will connect appear on the red the display … lead to … 10A and press the BLUE button one time Current (DC) (10 A max) and connect the black lead to … Hz VΩµA COM VΩµA COM VΩµA COM VΩµA COM 10 A COM 10 A COM VΩµA COM kHz or MHz Hz VΩµA COM MΩ or kΩ VΩµA COM mV or V VΩµA COM VΩµA COM µA µA A 10A A Ω Diode*** Frequency (DM-110 only) Resistance Voltage (AC) Voltage (DC) and press the BLUE button two times Hz Ω V and press the BLUE button one time V V mV or V * ** *** † Discharge capacitor before measurement. Tone sounds if the measured resistance is less than approximately 30 Ω. Diode voltage is shown if conducting, or “OL” is displayed for non-conductance. 3200 µA max for DM-100, 4000 µA max for DM-110. 11 Operation (cont’d) Auto Power Off The meter defaults to automatically turn itself off after a period of inactivity from the user. on the display indicates that the feature is enabled. The meter beeps approximately 15 seconds prior to turning off. The DM-100 turns off after 10 minutes, and the DM-110 turns off after 30 minutes. The present reading is held in memory. To turn the meter back on, press any button or turn the selector. The last reading will be displayed in the Hold mode. Press HOLD to return to normal operation. Disabling Auto Power Off To disable the Auto Power Off feature, press and hold the RANGE button while turning the meter on. Note: When the meter is turned off and back on with the selector, the meter returns to the default. However, the last value is not held in memory. 12 DM-100 • DM-110 Typical Measurements Voltage Measurement Current Measurement 13 Typical Measurements (cont’d) Resistance Measurement Continuity Check 14 DM-100 • DM-110 Typical Measurements (cont’d) Capacitance Measurement Diode Measurement Reverse Bias Forward Bias 15 Accuracy: DM-100 See the Specifications section for operating conditions and temperature coefficient. Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount) at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 80% relative humidity. For the DM-100, AC accuracy is specified for sine waves only. AC Current Range* Accuracy Burden Voltage 320.0 µA ± (2.2% + 0.5 µA) 1 V max 3200 µA ± (2.2% + 5 µA) 10 V max 10.00 A ± (2.5% + 0.05 A) 2 V max AC Voltage Range* 3.200 V 32.00 V 320.0 V 750 V Accuracy ± (1.50% + 0.005 V) ± (1.50% + 0.05 V) ± (1.50% + 0.5 V) ± (1.50% + 5 V) Frequency Range Input Impedance 40 to 500 Hz 10 MΩ, 100 pF inal DC Current Range* Accuracy Burden Voltage 320.0 µA ± (1.7% + 0.2 µA) 1 V max 3200 µA ± (1.7% + 2 µA) 10 V max 10.00 A ± (2% + 0.02 A) 2 V max * In autorange mode, ranges change at approximately 3200 counts. In manual range mode, range extends to 3400 counts. 16 DM-100 • DM-110 Accuracy: DM-100 (cont’d) DC Voltage Range* Accuracy Input Impedance 320.0 mV ± (0.5% + 0.2 mV) over 1000 MΩ 3.200 V ± (0.5% + 0.002 V) 32.00 V ± (0.5% + 0.02 V) 320.0 V ± (0.5% + 0.2 V) 1000 V ± (0.5% + 2 V) 10 MΩ Resistance Range* Accuracy 320.0 Ω ± (1% + 0.4 Ω) 3.200 kΩ ± (0.75% + 0.003 kΩ) 32.00 kΩ ± (0.75% + 0.03 kΩ) 320.0 kΩ ± (0.75% + 0.3 kΩ) 3.200 MΩ ± (1% + 0.003 MΩ) 32.00 MΩ ± (2% + 0.05 MΩ) Typical open circuit voltage: 1.5V * In autorange mode, ranges change at approximately 3200 counts. In manual range mode, range extends to 3400 counts. Diode Test Test Current (Typical): 1.5 mA Open Circuit Voltage: Less than 3.1 VDC Continuity Threshold: Tone sounds if the measured resistance is less than approximately 30 Ω. Response Time: Approximately 100 ms 17 Accuracy: DM-110 See the Specifications section for operating conditions and temperature coefficient. Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount) at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 80% relative humidity. For the DM-110, AC accuracy specification applies for sine waves at full scale and non-sine waves below half scale. • For crest factors from 1.4 to 3, add 1.5% to accuracy. • For crest factors from 3 to 4, add 3% to accuracy. AC Current Range Accuracy Burden Voltage 400.0 µA ± (2.2% + 0.5 µA) 1 V max 4000 µA ± (2.2% + 5 µA) 10 V max 10.00 A * ± (2.5% + 0.05 A) 2 V max * 10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off) AC Voltage Range Accuracy 400.0 mV unspecified 4.000 V ± (1.5% + 0.005 V) 40.00 V ± (1.3% + 0.05 V) 400.0 V ± (1.3% + 0.5 V) 750 V ± (1.3% + 5 V) Frequency Range Input Impedance 40 to 500 Hz 10 MΩ, 100 pF DC Current 18 Range Accuracy Burden Voltage 400.0 µA ± (1.7% + 0.2 µA) 1 V max 4000 µA ± (1.7% + 2 µA) 10 V max 10.00 A ± (2% + 0.02 A) 2 V max DM-100 • DM-110 Accuracy: DM-110 (cont’d) DC Voltage Range Accuracy Input Impedance 400.0 mV ± (0.5% + 0.2 mV) over 1000 MΩ 4.000 V ± (0.5% + 0.002 V) 40.00 V ± (0.5% + 0.02 V) 400.0 V ± (0.5% + 0.2 V) 1000 V ± (0.5% + 2 V) 10 MΩ Resistance Range Accuracy 400.0 Ω ± (1.0% + 0.3 Ω) 4.000 kΩ ± (0.75% + 0.002 kΩ) 40.00 kΩ ± (0.75% + 0.02 kΩ) 400.0 kΩ ± (0.75% + 0.2 kΩ) 4.000 MΩ ± (1.0% + 0.003 MΩ) 40.00 MΩ ± (1.5% + 0.05 MΩ) Typical open circuit voltage: 1.3V Capacitance Range Accuracy 4.000 nF ± (3.0% + 0.010 nF) 40.00 nF ± (3.0% + 0.10 nF) 400.0 nF ± (2.0% + 0.8 nF) 4.000 µF ± (2.0% + 0.008 µF) 40.00 µF ± (2.0% + 0.08 µF) 400.0 µF ± (2.0% + 0.8 µF) 4.000 mF* ± (5.0% + 0.020 mF) 40.00 mF* ± (5.0% + 0.20 mF) Accuracies are for film capacitors (capacitors with negligible dielectric absorption) *Accuracy specified for values less than 1/2 full scale. May have ripple within accuracy spec limits. 19 Accuracy: DM-110 (cont’d) Frequency Range Accuracy 4.000 kHz ± (0.01% + 0.001 kHz) Sensitivity 40.00 kHz ± (0.01% + 0.01 kHz) 400.0 kHz ± (0.01% + 0.1 kHz) 4.000 MHz ± (0.01% + 0.001 MHz) 250 mV rms 40.00 MHz ± (0.01% + 0.01 MHz) 1 V rms 100 mV rms * *Sensitivity is 1.5 V rms below 20 Hz The frequency measuring function has four selectable trigger levels. Use the range button to select. Diode Test Test Current (Typical): 1.5 mA Open Circuit Voltage: Less than 3.1 VDC Continuity Threshold: Tone sounds if the measured resistance is less than approximately 30 Ω. Response Time: Approximately 100 ms 20 DM-100 • DM-110 Specifications Display: DM-100: LCD (3200) and 70-segment bar graph In autorange mode, ranges change at approximately 3200 counts. In manual range mode, range extends to 3400 counts. DM-110: LCD (4000) and 82-segment bar graph Polarity: Automatic Sampling Rate: Numeric Display: 2 per second Bar Graph Display: 12 per second Temperature Coefficient: 0.15 x (Accuracy) per °C below 18 °C or above 28 °C Automatic Power-Off: DM-100: After 10 minutes of inactivity DM-110: After 30 minutes of inactivity Overload Protections: Volts: See Overvoltage Protection Categories Amps: A: 16A/500V HBC F Fuse, Interrupting Rating 10 kA, 1/4 x 1-1/4" Other Functions: 600 VDC/VAC RMS Overvoltage Protection Categories: VΩµA Terminal: Category III, 600 VAC and 600 VDC Category II, 1000 VAC and 1000 VDC Operating Conditions: 0 °C to 30 °C (32 °F to 86 °F), 0% to 80% Relative Humidity 30 °C to 40 °C (86 °F to 104 °F), 0% to 75% Relative Humidity 40 °C to 50 °C (104 °F to 122 °F), 0% to 45% Relative Humidity Indoor use only. Storage Conditions: -20 °C to 60 °C (-4 °F to 140 °F), 0% to 80% Relative Humidity Remove batteries Elevation: 2000 m (6500') maximum Pollution Degree: 2 Battery: DM-100: Two AAA 1.5-Volt batteries (NEDA 24A or IEC LR03) DM-110: One 9-Volt battery (NEDA 1604 or IEC 6F22) Battery Life: DM-100: Approximately 500 hours with alkaline DM-110: Approximately 300 hours with alkaline 21 Overvoltage Installation Categories These definitions were derived from the international safety standard for insulation coordination as it applies to measurement, control, and laboratory equipment. These overvoltage categories are explained in more detail by the International Electrotechnical Commission; refer to either of their publications: IEC 1010-1 or IEC 60664. Overvoltage Category I Signal level. Electronic and telecommunication equipment, or parts thereof. Some examples include transient-protected electronic circuits inside photocopiers and modems. Overvoltage Category II Local level. Appliances, portable equipment, and the circuits they are plugged into. Some examples include light fixtures, televisions, and long branch circuits. Overvoltage Category III Distribution level. Permanently installed machines and the circuits they are hard-wired to. Some examples include conveyor systems and the main circuit breaker panels of a building’s electrical system. Overvoltage Category IV Primary supply level. Overhead lines and other cable systems. Some examples include cables, meters, transformers, and other exterior equipment owned by the power utility. 22 DM-100 • DM-110 Maintenance • Do not attempt to repair this unit. It contains no user-serviceable parts. • Do not expose the unit to extremes in temperature or high humidity. See Specifications. Failure to observe these precautions can result in injury and can damage the unit. Battery and Fuse Replacement Before opening the case, remove the test leads from the circuit and shut off the unit. Failure to observe this warning can result in severe injury or death. The fuse is an integral part of the overvoltage protection. When fuse replacement is necessary, see Specifications for the correct type, size, and capacity. Using any other type of fuse will void the overvoltage protection rating of the unit. Failure to observe this warning can result in severe injury or death. 1. Disconnect the unit from the circuit. Turn the unit OFF. 2. Remove the screws from the back cover. 3. Remove the back cover. 4. Lift battery holder out of back cover. 5. Replace the batteries (observe polarity) and/or fuse. 6. Replace the cover, screws and stand. Cleaning Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. 23 Greenlee Textron / Subsidiary of Textron Inc. 24 4455 Boeing Dr., Rockford, IL 61109-2988 815/397-7070 DM-100 • DM-110 Descripción Los Multímetros digitales modelos DM-100 y DM-110 de Greenlee son instrumentos de verificación capaces de efectuar los siguientes tipos de mediciones: tensión alterna y continua, corriente alterna y continua, y resistencia. Estas unidades son de bolsillo y caben perfectamente en la palma de la mano. También sirven para verificar diodos y continuidad. Además, el modelo DM-110 es capaz de efectuar mediciones de frecuencia y capacitancia. El modelo DM-110 es un multímetro de lectura de valores eficaces reales. Ambos modelos incluyen la función de apagado automático para conservar la vida útil de las pilas, además de un gráfico de barras, el cual es más sensible que la pantalla numérica – útil para detectar fallas en contatos, potenciómetros y picos de señal. Acerca de la seguridad Es fundamental observar métodos seguros al utilizar y dar mantenimiento a las herramientas y equipo Greenlee. Este manual de instrucciones y todas las marcas que ostenta la herramienta le ofrecen la información necesaria para evitar riesgos y hábitos poco seguros relacionados con su uso. Siga toda la información sobre seguridad que se proporciona. Propósito de este manual Este manual de instrucciones tiene como propósito familiarizar a todo el personal con los procedimientos de operación y mantenimiento seguros para los Multímetros digitales modelos DM-100 y DM-110 de Greenlee. Manténgalo siempre al alcance de todo el personal. Puede obtener copias adicionales de manera gratuita, previa solicitud. Greenlee y son marcas registradas de Greenlee Textron. CONSERVE ESTE MANUAL 25 Importante Información sobre Seguridad SÍMBOLO DE ALERTA SOBRE SEGURIDAD Este símbolo se utiliza para indicar un riesgo o práctica poco segura que podría ocasionar lesiones o daños materiales. Cada uno de los siguientes términos denota la gravedad del riesgo. El mensaje que sigue a dichos términos le indica cómo puede evitar o prevenir ese riesgo. Peligros inmediatos que, de no evitarse, OCASIONARÁN graves lesiones o incluso la muerte. Peligros que, de no evitarse, PODRÍAN OCASIONAR graves lesiones o incluso la muerte. Peligro o prácticas peligrosas que, de no evitarse, PUEDEN OCASIONAR lesiones o daños materiales. Lea y entienda este documento antes de manejar esta herramienta o darle mantenimiento. Utilizarla sin comprender cómo manejarla de manera segura podría ocasionar un accidente y, como resultado de éste, graves lesiones o incluso la muerte. 26 DM-100 • DM-110 Importante Información sobre Seguridad Peligro de electrocución: El contacto con circuitos activados puede ocasionar graves lesiones o incluso la muerte. Peligro de electrocución e incendio: • No exponga esta unidad ni a la lluvia ni a la humedad. • No utilice esta unidad si se encuentra mojada o dañada. • Utilice cables de prueba y accesorios que sean apropiados para la aplicación que se va a realizar. Consulte la información sobre categoría y voltaje nominal del cable de prueba o el accesorio. • Revise minuciosamente los cables de prueba o el accesorio, antes de utilizarlos. Deberán estar limpios y secos, y su forro aislante deberá hallarse en buenas condiciones. • Utilícela únicamente para el propósito para el que ha sido diseñada por el fabricante, tal como se describe en este manual. Cualquier otro uso puede menoscabar la protección proporcionada por la unidad. De no observarse estas advertencias pueden sufrirse graves lesiones o incluso la muerte. Peligro de electrocución: • No aplique más del voltaje nominal entre dos terminales de entrada cualesquiera, o entre una terminal de entrada cualquiera y una conexión a tierra. • No toque las puntas de los cables de prueba ni ninguna parte del accesorio que carezca de forro aislante. De no observarse estas advertencias pueden sufrirse graves lesiones o incluso la muerte. • No haga funcionar esta unidad con la caja abierta. • Antes de abrir la caja, retire del circuito los cables de prueba y apague la unidad. De no observarse estas advertencias pueden sufrirse graves lesiones o incluso la muerte. 27 Importante Información sobre Seguridad El fusible es una parte integral para la protección contra sobretensión. Cuando sea necesario reemplazarlo, consulte la sección “Especificaciones” para saber qué tipo, tamaño y capacidad debe tener. Utilizar cualquier otro tipo de fusible anulará la clasificación de protección de sobretensión de la unidad. De no observarse esta advertencia pueden sufrirse graves lesiones o incluso la muerte. No cambie la función de medición mientras los cables de prueba estén conectados a un componente o circuito. De no observarse esta advertencia podrían sufrirse lesiones o daños a la unidad. • No intente reparar esta unidad, ya que contiene partes que deben recibir mantenimiento por parte de un profesional. • No exponga la unidad a ambientes de temperatura extrema o altos niveles de humedad; Consulte las Especificaciones. De no observarse estas precauciones podrían sufrirse lesiones o daños a la unidad. • A menos que vaya a medir tensión, corriente o frecuencia, apague y bloquee la energía. Asegúrese de que todos los condensadores estén totalmente sin carga. No debe haber tensión alguna. • Coloque el interruptor de selección y conecte los cables de prueba de modo que correspondan al tipo de medición que se desea efectuar. Si se colocan o se conectan incorrectamente puede quemarse un fusible. • Al utilizar esta unidad cerca de equipo que genere interferencia electromagnética quizá se obtenga una lectura inexacta e inestable. 28 DM-100 • DM-110 Identificación 1. Pantalla Pantalla de cristal líquido (LCD) de 4 dígitos y gráfico de barras 2. Botones Selectores de funciones Véase la explicación pertinente en la sección “Cómo utilizar las distintas funciones” 3. Selector Selecciona una función o apaga (OFF) la unidad 4. Hz VΩµA Terminal de señal de entrada positiva para todas las mediciones excepto corriente de 10A ( Hz en modelo DM-110 únicamente) 5. COM Terminal de entrada a tierra, común o negativa, para todo tipo de mediciones 6. 10A Terminal de entrada positiva para mediciones de corriente de hasta 10 A 1 2 3 5 4 6 Símbolos en la unidad Advertencia—Lea el manual de instrucciones Pila Fusible Doble forro aislante 29 Iconos de la pantalla Ambos modelos Modelo DM-100 únicamente 7. Se selecciona Medición manual de escala 8. Se selecciona medición de CC 9. 22. Se activa la función “Hold” (Retención de datos en pantalla) Se selecciona medición de CA Modelo DM-110 únicamente 10. Pila baja 23. 11. APO Se activa la función Apagado automático Se activa la función “Hold” (Retención de datos en pantalla) 12. Diodo 24. n nano (10-9) 13. Se selecciona Continuidad 25. µ Micro (10-6) 14. µ Micro (10-6) 26. m Mili (10-3) Faradios 15. m Mili (10 ) 27. F 16. V Voltios Nota: Los iconos sin identificar no se utilizan en estos modelos. -3 17. A Amperios 18. M Mega (106) 19. k Kilo (103) 20. Ω Ohmios 21. Hz Herzios (frecuencia o ciclos por segundo) Pantalla del modelo DM-110 7 23 11 12 13 14,15,16,17 24,25,26,27 18,19,20,21 10 8 9 12 13 Pantalla del modelo DM-100 22 11 7 8 9 10 30 14,15,16,17 18,19,20,21 DM-100 • DM-110 Medición de corriente alterna Las mediciones de corriente alterna generalmente se muestran como valores eficaces (RMS o root mean squared). Existen dos métodos de medición de corriente alterna: calibrados para responder al valor eficaz medio y a una lectura de valores eficaces reales. El método calibrado para responder al valor eficaz medio toma el valor medio de la señal de entrada, la multiplica por 1,11 y muestra el resultado. El resultado es exacto si la señal de entrada es una onda sinusoidal pura. El modelo DM-100 de Greenlee es un multímetro que responde al valor medio. El método de lectura de valores eficaces reales utiliza un circuito interno para leer el valor eficaz real. Este método es exacto, dentro de las limitaciones de factor de cresta especificadas, independientemente del tipo de señal de entrada, ya sea una onda sinusoidal pura, rectangular, triangular, media onda o señal con armónicas. La capacidad para leer valores eficaces reales brinda una mayor versatilidad de medición. La unidad modelo DM-110 de Greenlee es un medidor de valores eficaces reales. La tabla de Formas de onda y Factores de cresta muestra algunas de las señales de CA y sus valores eficaces reales más comunes. Formas de onda y Factores de cresta Forma de onda Valor eficaz real Valor medio *Factor de cresta (ξ) 100 90 1,414 100 100 1 100 87 1,73 100 64 2 * El factor de cresta es el cociente de un valor máximo en relación con el valor eficaz; está representado por la letra griega ξ. 31 Cómo utilizar las distintas funciones Botón Azul (sin etiqueta) Oprima momentáneamente para: Notas: • seleccionar CA o CC para medir corriente o tensión • seleccionar funciones secundarias RANGE Oprima y mantenga oprimido para cambiar de selección manual de escala a selección automática de escala. PWR RST (DM-100 únicamente) volver a encender luego de apagado automático Hz (DM-110 únicamente) ver la frecuencia mientras mide tensión de CA o corriente Mientras se encuentre en el modo Hz, el botón RANGE no cambia la escala de frecuencia. Sin embargo, cambia la escala de tensión. retener un valor medido para todas las funciones. Oprima nuevamente para regresar al modo normal Se efectuan las conversiones pero la pantalla no las actualiza. HOLD 32 • cambiar de selección automática de escala al modo de selección manual de escala • seleccionar escalas al utilizar el modo de selección manual de escala • seleccionar niveles de activación al medir frecuencia DM-100 • DM-110 Operación 1. Consulte la Tabla de valores. Coloque el interruptor de selección en el valor apropiado, oprima el botón azul sin etiqueta para seleccionar la medición deseada y conecte los cables de prueba al medidor. Empiece por el ajuste de medición más alto. 2. Consulte la sección “Mediciones más comunes” en relación con las instrucciones específicas para cada tipo de medición. 3. Pruebe la unidad en un circuito o componente que se sabe está funcionando perfectamente. • Si no funciona como debería en un circuito que se sabe está funcionando perfectamente, reemplace la pila y/o el fusible. • Si sigue sin funcionar como debería, devuélvala a Greenlee a fin de que sea reparada. Vea la dirección bajo Garantía. 4. Anote la lectura del circuito o componente que se está verificando. Si la resolución de la lectura no es satisfactoria, desconecte el medidor del circuito y cambie a la siguiente escala inferior si está utilizando la selección manual de escala. 33 Operación (continuación) Tabla de valores Para medir esto... coloque en este símbolo... Capacitancia* (DM-110 únicamente) Continuidad** Corriente (CA)† Corriente (CC)† Ω y oprima el botón AZUL una sola vez µA y oprima el botón AZUL una sola vez µA conecte este icono el cable aparecerá en de prueba la pantalla… rojo a… F Hz VΩµA y conecte el cable de prueba negro a… COM VΩµA COM VΩµA COM VΩµA COM 10 A COM 10 A COM VΩµA COM COM MΩ o kΩ Hz VΩµA VΩµA mV o V VΩµA COM VΩµA COM µA µA Corriente (CA) (10 A máx) Corriente (CC) (10 A máx) Diodo*** Frecuencia (DM-110 únicamente) Resistencia Tensión (CA) Tensión (CC) 10A y oprima el botón AZUL una sola vez 10A A A Ω y oprima el botón AZUL dos veces Hz Ω V y oprima el botón AZUL una sola vez V V kHz o MHz mV o V * Descargue el condensador antes de efectuar una medición. ** El tono suena si la resistencia medida es menor a 30 Ω aproximadamente. *** Se indica la tensión de diodo si se encuentra en estado de conducción, o se muestra “OL” si se encuentra en estado aislante. † 3200 µA máx para DM-100, 4000 µA máx para DM-110. 34 COM DM-100 • DM-110 Operación (continuación) Apagado automático El medidor conmutará por defecto para apagarse automáticamente luego de un período de inactividad por parte del usuario. en la pantalla indica que la función se ha activado. El medidor emite un tono audible por espacio de 15 segundos antes de apagarse. El modelo DM-100 se apaga luego de 10 minutos y el modelo DM-110 se apaga luego de 30 minutos. La lectura actual se retendrá en la memoria. Para volver a encender el medidor, oprima cualquier botón o gire el interruptor de selección. La última lectura se mostrará en el modo Hold (Retención de datos en pantalla). Oprima HOLD para regresar a la operación normal. Desactivación de la función de Apagado automático Para desactivar la función de Apagado automático, oprima y mantenga oprimido el botón RANGE al tiempo que enciende la unidad. Nota: Cuando el medidor se apaga y se vuelve a encender con el interruptor de selección, la unidad regresará a sus valores predeterminados. Sin embargo, la memoria no retendrá el último valor. 35 Mediciones más comunes Medición de tensión Medición de corriente 36 DM-100 • DM-110 Mediciones más comunes (continuación) Medición de resistencia Verificación de continuidad 37 Mediciones más comunes (continuación) Medición de capacitancia Verificación de diodo Polarización inversa Polarización directa 38 DM-100 • DM-110 Precisión: DM-100 Consulte la sección “Especificaciones” en relación con las condiciones de operación y el coeficiente de temperatura. La precisión se especifica de la siguiente manera: ± (un porcentaje de la lectura + una cantidad fija) a 23°C a ± 5°C (73,4°F ± 9°F), 0% a 80% de humedad relativa. Para el modelo DM-100, la precisión de CA se especifica para ondas sinusoidales únicamente. Corriente alterna (CA) Escala* Precisión Tensión de carga 320,0 µA ± (2,2% + 0,5 µA) 1V máx 3200 µA ± (2,2% + 5 µA) 10V máx 10,00 A ± (2,5% + 0,05 A) 2V máx Tensión alterna (CA) Escala* Precisión 3,200V ± (1,50% + 0,005V) 32,00V ± (1,50% + 0,05V) 320,0V ± (1,50% + 0,5V) 750V ± (1,50% + 5V) Escala de frecuencia Impedancia de entrada 40 a 500 Hz 10 MΩ, 100 pF inal Corriente continua (CC) Escala* Precisión 320,0 µA ± (1,7% + 0,2 µA) Tensión de carga 1V máx 3200 µA ± (1,7% + 2 µA) 10V máx 10,00 A ± (2% + 0,02 A) 2V máx * En el modo de selección automática de escala, la escala cambia a aproximadamente 3.200 puntos. En el modo de selección manual de escala, la escala se extiende a 3.400 puntos. 39 Precisión: DM-100 (continuación) Tensión continua (CC) Escala* Precisión Impedancia de entrada 320,0 mV ± (0,5% + 0,2 mV) superior a 1000 MΩ 3,200V ± (0,5% + 0,002V) 32,00V ± (0,5% + 0,02V) 320,0V ± (0,5% + 0,2V) 1000V ± (0,5% + 2V) 10 MΩ Resistencia Escala* Precisión 320,0 Ω ± (1% + 0,4 Ω) 3,200 kΩ ± (0,75% + 0,003 kΩ) 32,00 kΩ ± (0,75% + 0,03 kΩ) 320,0 kΩ ± (0,75% + 0,3 kΩ) 3,200 MΩ ± (1% + 0,003 MΩ) 32,00 MΩ ± (2% + 0,05 MΩ) Tensión de circuito abierto típica: 1,5V * En el modo de selección automática de escala, la escala cambia a aproximadamente 3.200 puntos. En el modo de selección manual de escala, la escala se extiende a 3.400 puntos. Verificación de diodo Corriente de prueba (Típica): 1,5 mA Tensión de circuito abierto: Menor de 3,1V CC Continuidad Umbral: El tono suena si la resistencia medida es menor a 30 Ω aproximadamente. Tiempo de respuesta: Aproximadamente 100 ms. 40 DM-100 • DM-110 Precisión: DM-110 Consulte la sección “Especificaciones” en relación con las condiciones de operación y el coeficiente de temperatura. La precisión se especifica de la siguiente manera: ± (un porcentaje de la lectura + una cantidad fija) a 23°C a ± 5°C (73,4°F ± 9°F), 0% a 80% de humedad relativa. Para el modelo DM-110, la especificación sobre precisión de CA aplica a ondas sinusoidales a plena escala y ondas no sinusoidales por debajo de la escala media. • Para factores de cresta de 1,4 a 3, agregue 1,5% a la precisión. • Para factores de cresta de 3 a 4, agregue 3% a la precisión. Corriente alterna (CA) Escala Precisión Tensión de carga 400,0 µA ± (2,2% + 0,5 µA) 1V máx 4000 µA ± (2,2% + 5 µA) 10V máx 10,00 A* ± (2,5% + 0,05 A) 2V máx * 10 A continuos; 20 A máximo (Régimen de trabajo: 30 segundos encendido, 5 minutos apagado) Tensión alterna (CA) Escala Precisión 400,0 mV no especificado 4,000V ± (1,5% + 0,005V) 40,00V ± (1,3% + 0,05V) 400,0V ± (1,3% + 0,5V) 750V ± (1,3% + 5V) Escala de frecuencia Impedancia de entrada 40 a 500 Hz 10 MΩ, 100 pF Corriente continua (CC) Escala Precisión 400,0 µA ± (1,7% + 0,2 µA) Tensión de carga 1V máx 4000 µA ± (1,7% + 2 µA) 10V máx 10,00 A ± (2% + 0,02 A) 2V máx 41 Precisión: DM-110 (continuación) Tensión continua (CC) Escala Precisión Impedancia de entrada 400,0 mV ± (0,5% + 0,2 mV) superior a 1000 MΩ 4,000V ± (0,5% + 0,002V) 40,00V ± (0,5% + 0,02V) 400,0V ± (0,5% + 0,2V) 1000V ± (0,5% + 2V) 10 MΩ Resistencia Escala Precisión 400,0 Ω ± (1,0% + 0,3 Ω) 4,000 kΩ ± (0,75% + 0,002 kΩ) 40,00 kΩ ± (0,75% + 0,02 kΩ) 400,0 kΩ ± (0,75% + 0,2 kΩ) 4,000 MΩ ± (1,0% + 0,003 MΩ) 40,00 MΩ ± (1,5% + 0,05 MΩ) Tensión de circuito abierto típica: 1,3V Capacitancia Escala Precisión 4,000 nF ± (3,0% + 0,010 nF) 40,00 nF ± (3,0% + 0,10 nF) 400,0 nF ± (2,0% + 0,8 nF) 4,000 µF ± (2,0% + 0,008 µF) 40,00 µF ± (2,0% + 0,08 µF) 400,0 µF ± (2,0% + 0,8 µF) 4,000 mF* ± (5,0% + 0,020 mF) 40,00 mF* ± (5,0% + 0,20 mF) Estas precisiones son para los condensadores de película (condensadores con absorción dieléctrica inapreciable) *Precisión especificada para valores menores de media escala completa. Puede haber fluctuacion dentro de los límites de precisión específicos. 42 DM-100 • DM-110 Precisión: DM-110 (continuación) Frecuencia Escala Precisión 4,000 kHz ± (0,01% + 0,001 kHz) Sensitividad 40,00 kHz ± (0,01% + 0,01 kHz) 400,0 kHz ± (0,01% + 0,1 kHz) 4,000 MHz ± (0,01% + 0,001 MHz) 250 mV rms * 40,00 MHz ± (0,01% + 0,01 MHz) 1V rms 100 mV rms * *La sensitividad es 1,5V rms por debajo de 20 Hz Con la función de medición de frecuencia se puede seleccionar entre cuatro niveles de activación. Utilice el botón Range para seleccionar el deseado. Verificación de diodo Corriente de prueba (Típica): 1,5 mA Tensión de circuito abierto: Menor de 3,1V CC Continuidad Umbral: El tono suena si la resistencia medida es menor a 30 Ω aproximadamente. Tiempo de respuesta: Aproximadamente 100 ms. 43 Especificaciones Pantalla: DM-100: cristal líquido (LCD) (3200) y gráfico de barras de 70 segmentos En el modo de selección automática, la escala cambia a aproximadamente 3200 puntos. En el modo de selección manual de escala, la escala se extiende a 3400 puntos. DM-110: cristal líquido (LCD) (4000) y gráfico de barras de 82 segmentos Polaridad: Automática Frecuencia de muestreo: Pantalla numérica: 2 por segundo Pantalla de gráfico de barras: 12 por segundo Coeficiente de temperatura: 0,15 x (precisión) por °C menor de 18°C o mayor de 28°C Apagado automático: DM-100: Después de 10 minutos de inactividad DM-110: Después de 30 minutos de inactividad Protecciones contra sobrecarga: Voltios: Consulte las Categorías de protección de sobretensión Amps: A: Fusible16A/500V HBC F, condiciones de corte 10 kA, 1/4 x 1-1/4 pulg. Otras funciones: 600V CC/V CA eficaces Categorías de protección de sobretensión: Terminal VΩµA: Categoría III, 600V CA y 600V CC Categoría II, 1000V CA y 1000V CC Condiciones de operación: 0°C a 30°C (32°F a 86°F), 0% a 80% de humedad relativa 30°C a 40°C (86°F a 104°F), 0% a 75% de humedad relativa 40°C a 50°C (104°F a 122°F), 0% a 45% de humedad relativa Uso en interiores únicamente. Condiciones de almacenamiento: -20°C a 60°C (-4°F a 140°F), 0% a 80% de humedad relativa Retire las pilas Altura: 2.000 m (6.500 pies) máximo Grado de contaminación: 2 Pila: DM-100: Dos pilas AAA de 1,5V (NEDA 24A o IEC LR03) DM-110: Pila de 9 voltios (NEDA 1604 o IEC 6F22) Vida útil de la pila: DM-100: Aproximadamente 500 horas con pilas alcalinas DM-110: Aproximadamente 300 horas con pilas alcalinas 44 DM-100 • DM-110 Categorías de instalación de sobretensión Las siguientes definiciones proceden de la norma de seguridad internacional sobre la coordinación de aislamientos tal y como se aplica a equipos de medición, control y laboratorio. En las publicaciones IEC 1010-1 y IEC 60664 de la International Electrotechnical Commission (Comisión Electrotécnica Internacional) se detallan más a fondo estas categorías de sobretensión. Categoría de sobretensión I Nivel de señal. Equipo electrónico y de telecomunicaciones, o partes del mismo. Como ejemplo pueden citarse los circuitos electrónicos protegidos contra tensiones momentáneas dentro de fotocopiadores y modems. Categoría de sobretensión II Nivel local. Aparatos eléctricos, equipo portátil, y los circuitos a los que están conectados. Como ejemplo pueden citarse dispositivos de iluminación, televisores y circuitos de rama larga. Categoría de sobretensión III Nivel de distribución. Máquinas instaladas permanentemente y los circuitos a los que están cableados. Como ejemplo pueden citarse sistemas conductores y los paneles del interruptor automático principal del sistema eléctrico de un edificio. Categoría de sobretensión IV Nivel de abastecimiento primario. Líneas aéreas y otros sistemas de cable. Como ejemplo pueden citarse cables, medidores, transformadores y cualquier otro equipo exterior perteneciente a la empresa de servicio eléctrico. 45 Mantenimiento • No intente reparar esta unidad, ya que contiene partes que deben recibir mantenimiento por parte de un profesional. • No exponga la unidad a ambientes de temperatura extrema o altos niveles de humedad; Consulte las Especificaciones. De no observarse estas precauciones podrían sufrirse lesiones o daños a la unidad. Cómo reemplazar la pila y el fusible Antes de abrir la caja, retire del circuito los cables de prueba y apague la unidad. De no observarse esta advertencia pueden sufrirse graves lesiones o incluso la muerte. El fusible es una parte integral para la protección contra sobretensión. Cuando sea necesario reemplazarlo, consulte la sección “Especificaciones” para saber qué tipo, tamaño y capacidad debe tener. Utilizar cualquier otro tipo de fusible anulará la clasificación de protección de sobretensión de la unidad. De no observarse esta advertencia pueden sufrirse graves lesiones o incluso la muerte. 1. Desconecte la unidad del circuito. Apague (OFF) la unidad. 2. Retire los tornillos de la tapa posterior. 3. Retire la tapa posterior. 4. Levante el compartimiento de la pila de la tapa posterior. 5. Reemplace las pilas (fíjese en la polaridad) y/o el (los) fusible(s). 6. Vuelva a colocar la tapa, los tornillos y el apoyo. Limpieza Limpie periódicamente la caja utilizando un paño húmedo y detergente suave; no utilice abrasivos ni solventes. 46 DM-100 • DM-110 Greenlee Textron / Subsidiary of Textron Inc. 4455 Boeing Dr., Rockford, IL 61109-2988 815/397-7070 47 Greenlee Textron / Subsidiary of Textron Inc. 4455 Boeing Dr., Rockford, IL 61109-2988 815/397-7070 DM-100 • DM-110 Description Les multimètres numériques DM-100 et DM-110 sont des appareils de vérification portables dotés des capacités de mesure suivantes : la tension c.a. et c.c., le courant c.a. et c.c. et la résistance. Ils contrôlent aussi les diodes et vérifient la continuité. En outre, le DM-110 mesure la fréquence et la capacité. Le DM-110 est un multimètre à lecture RMS véritable. Les deux modèles sont équipés d’une fonction de mise hors tension automatique qui économise les piles et d’un affichage graphique plus sensible que l’affichage numérique — utile pour détecter les contacts défectueux, les clics de potentiomètre et les brusques variations de signal. Sécurité Lors de l’utilisation et de l’entretien des outils et des équipements de Greenlee, votre sécurité est une priorité. Ce manuel d’instructions et toute étiquette sur l’outil fournit des informations permettant d’éviter des dangers ou des manipulations dangereuses liées à l’utilisation de cet outil. Suivre toutes les consignes de sécurité indiquées. Dessein Ce manuel d’instructions est conçu pour que le personnel puisse se familiariser avec le fonctionnement et les procédures d’entretien sûres des multimètres numériques DM-100 et DM-110 de Greenlee. Mettre ce manuel à la disposition de tous les employés. On peut obtenir des exemplaires gratuits sur simple demande. Greenlee et sont des marques déposées de Greenlee Textron. CONSERVER CE MANUEL 49 Consignes de sécurité importantes SYMBOLE D’AVERTISSEMENT Ce symbole met en garde contre les risques et les manipulations dangereuses pouvant entraîner des blessures ou l’endommagement du matériel. Le mot indicateur, défini cidessous, indique la gravité du danger. Le message qui suit le mot indicateur indique comment empêcher le danger. Danger immédiat qui, s’il n’est pas pris en considération ENTRAINERA des blessures graves, voire mortelles. Danger qui, s’il n’est pas pris en considération, POURRAIT entraîner des blessures graves, voire mortelles. Dangers ou manipulations dangereuses qui, s’ils ne sont pas pris en considération, POURRAIENT EVENTUELLEMENT entraîner des dommages à la propriété ou causer des blessures. Lire attentivement et bien comprendre cette documentation avant d’utiliser ou de procéder à l’entretien de cet équipement. Négliger de comprendre comment utiliser cet outil en toute sécurité peut provoquer un accident et entraîner des blessures graves, voire mortelles. 50 DM-100 • DM-110 Consignes de sécurité importantes Risques de décharge électrique : Un contact avec des circuits sous tension peut entraîner des blessures graves, voire mortelles. Risques de décharge électrique et d’incendie : • Ne pas exposer cet appareil à la pluie ou à l’humidité. • Ne pas utiliser cet appareil s’il est mouillé ou endommagé. • Utiliser des fils d’essai ou des accessoires conformes à l’application. Consulter la catégorie et la tension nominale du fil d’essai ou de l’accessoire. • Vérifier les fils d’essai ou l’accessoire avant de les utiliser. La pièce (ou les pièces) doi(ven)t être propre(s) et sèche(s) et l’isolation en bon état. • Utiliser cet appareil uniquement dans le but pour lequel il a été conçu, tel que décrit dans ce manuel. Toute autre utilisation peut altérer le système de protection de cet appareil. L’inobservation de ces consignes peut entraîner des blessures graves, voire mortelles. Risques de décharge électrique : • Ne pas appliquer plus que la tension nominale entre deux bornes d’entrée, ou entre une borne d’entrée et une prise de terre. • Ne pas entrer en contact avec les extrémités des fils d’essai ou avec toute autre partie non isolée de l’accessoire. L’inobservation de ces consignes peut entraîner des blessures graves, voire mortelles. • Ne pas utiliser lorsque le boîtier est ouvert. • Avant d’ouvrir le boîtier, retirer les fils d’essai du circuit et mettre l’appareil hors tension. L’inobservation de ces consignes peut entraîner des blessures graves, voire mortelles. 51 Consignes de sécurité importantes Le fusible est une partie intégrante du système de protection de surtension. Lorsqu’un fusible doit être remplacé, consulter les spécifications pour connaître le type, la taille et la capacité requis. L’utilisation de tout autre type de fusible annule le calibrage de protection contre la surtension de l’appareil. L’inobservation de cette consigne peut entraîner des blessures graves, voire mortelles. Ne pas modifier la fonction de mesure pendant que les fils d’essai sont connectés à un composant ou à un circuit. L’inobservation de cette consigne peut endommager l’appareil et entraîner des blessures. • Ne pas tenter de réparer cet appareil. Il ne comporte aucune pièce pouvant être réparée. • Ne pas exposer l’appareil à des températures ou à une humidité extrêmes. Voir les spécifications. L’inobservation de ces consignes peut endommager l’appareil et entraîner des blessures. • Sauf si l’on mesure la tension, le courant ou la fréquence, mettre hors tension et couper la source d’alimentation. S’assurer que tous les condensateurs sont déchargés. Aucune tension ne doit être présente. • Régler le sélecteur et connecter les fils d’essai pour qu’ils correspondent à la mesure voulue. Des réglages ou des connexions incorrects peuvent faire sauter les fusibles. • L’utilisation de cet appareil à proximité d’équipements qui génèrent des interférences électromagnétiques peut produire des lectures instables ou erronées. 52 DM-100 • DM-110 Identification 1. Affichage Graphique à barres sur cristaux liquides avec 4 chiffres 2. Boutons de fonction Voir les explications de la section Utilisation des fonctions 3. Sélecteur Sélectionne une fonction ou met hors tension (OFF). 4. Hz VΩµA Borne d’entrée positive pour toutes les mesures sauf pour le courant de 10 A ( Hz sur le DM-110 uniquement) 5. COM Borne d’entrée négative, commune ou reliée à la terre, pour toutes les mesures 6. 10A Borne d’entrée positive pour les mesures de courant allant jusqu’à 10 A 1 2 3 5 4 6 Symboles apparaissant sur l’appareil Avertissement — Lire le manuel d’instructions Pile Fusible Isolation double 53 Icônes d’affichage DM-100 uniquement Les deux modèles 7. Plage manuelle est sélectionnée 22. 8. Mesure du c.c. sélectionnée DM-110 uniquement 9. Mesure du c.a. sélectionnée 23. Fonction de maintien activée 10. Pile faible ou déchargée 24. n nano (10-9) 11. APO Mise hors tension automatique activée 25. µ Micro (10-6) 12. Diode 26. m Milli (10-3) 27. F Farads 13. Continuité est sélectionné 14. µ Micro (10-6) 15. m Milli (10-3) 16. V Volts 17. A Ampères 18. M Méga (106) 19. k Kilo (103) 20. Ω Ohms 21. Hz Hertz (fréquence ou cycles par seconde) Fonction de maintien activée Remarque : Les icônes non identifiées ne sont pas utilisées sur ces modèles. Affichage du DM-110 7 23 11 12 13 14,15,16,17 24,25,26,27 18,19,20,21 10 8 9 12 13 Affichage du DM-100 22 11 7 8 9 10 54 14,15,16,17 18,19,20,21 DM-100 • DM-110 Mesure du c.a. En règle générale, les mesures du c.a. sont affichées en valeurs RMS (moyenne quadratique). Les deux méthodes de mesure du c.a. sont calibrées pour une réponse RMS moyenne et pour une lecture RMS véritable. La méthode par valeur moyenne consiste à mesurer la valeur moyenne du signal d’entrée, à la multiplier par 1,11 et à afficher le résultat. Cette méthode est précise si le signal d’entrée est une onde sinusoïdale pure. Le Greenlee DM-100 mesure la réponse moyenne. La méthode par mesure de la valeur efficace vraie (RMS) fait appel à un circuit interne permettant de mesurer directement la valeur efficace. Cette méthode est précise dans les limites du facteur de crête spécifiées, que le signal soit une onde pure, une onde carrée, une onde triangulaire, une demi-onde ou un signal comportant des harmoniques. Les appareils à mesure de valeur efficace vraie (RMS) sont beaucoup plus polyvalents que les appareils conventionnels. Le multimètre DM-110 de Greenlee lit la valeur RMS véritable. Le tableau des formes d’onde et facteurs de crête fournit les valeurs efficaces des signaux alternatifs courants. Formes d’ondes et facteurs de crêtes Forme d’onde Valeur RMS Valeur moyenne Facteur de crête* (ξ) 100 90 1,414 100 100 1 100 87 1,73 100 64 2 * Le facteur de crête correspond au ratio de la valeur de crête par rapport à la valeur RMS ; il est représenté par la lettre grecque ξ. 55 Utilisation des fonctions Bouton Appuyer quelques secondes pour : Bleu (pas d’étiquette) • sélectionner c.a. ou c.c. lors de la mesure du courant ou de la tension Remarques : • sélectionner les fonctions secondaires • changer à partir de la mesure automatique de la plage à plage manuelle • sélectionner les plages lors de l’utilisation du mode de plage manuelle • sélectionner les niveaux de déclenchement pendant la fréquence de mesure PWR RST (DM-100 uniquement) remettre en marche après l’extinction auto Hz (DM-110 uniquement) visionner la fréquence pendant la mesure de la tension c.a. ou du courant En mode Hz, le bouton RANGE ne change pas la plage de fréquence. Toutefois, il change la plage de tension. maintenir une valeur mesurée pour toutes les fonctions. Presser de nouveau pour revenir au mode normal Les conversions sont effectuées mais l’affichage ne se met pas à jour. HOLD 56 Presser et maintenir pour changer la sélection de plage de manuelle à automatique. RANGE DM-100 • DM-110 Utilisation 1. Se reporter au tableau des réglages. Régler le sélecteur au réglage approprié, appuyer sur le bouton bleu sans étiquette pour sélectionner la mesure voulue et connecter les fils d’essai au multimètre. Commencer avec le réglage de mesures le plus élevé. 2. Se reporter aux mesures types pour obtenir des instructions de mesure spécifiques. 3. Vérifier l’appareil sur un circuit ou sur un composant connu. • Si l’appareil ne fonctionne pas comme prévu sur un circuit dont le fonctionnement est connu, remplacer la pile et/ou le fusible. • Si l’appareil ne fonctionne toujours pas comme prévu, le renvoyer à Greenlee pour qu’il soit réparé. Voir l’adresse indiquée sous la garantie. 4. Lire le circuit ou le composant à vérifier. Si la résolution n’est pas satisfaisante, retirer le multimètre du circuit et passer à la plage immédiatement moins élevée à l’aide de la sélection manuelle de la plage. 57 Utilisation (suite) Tableau des réglages Pour mesurer cela ... régler le sélecteur à ce symbole … F Capacité* (DM-110 uniquement) Continuité** Courant (c.a.)† Courant (c.c.)† cette icône connecter apparaît sur le fil sur l’écran … rouge à… Ω et appuyer sur le bouton BLEU une fois µA et appuyer sur le bouton BLEU une fois µA et connecter le fil noir à ... Hz VΩµA COM VΩµA COM VΩµA COM VΩµA COM 10 A COM 10 A COM VΩµA COM COM MΩ ou kΩ Hz VΩµA VΩµA mV ou V VΩµA COM VΩµA COM µA µA Courant (c.a.) (10 A max.) Courant (c.c.) (10 A max.) Diode*** Fréquence (DM-110 uniquement) Résistance Tension (c.a.) Tension (c.c.) 10A et appuyer sur le bouton BLEU une fois 10A A A Ω et appuyer sur le bouton BLEU deux fois Hz Ω V et appuyer sur le bouton BLEU une fois V V kHz ou MHz mV ou V * Décharger le condensateur avant de mesurer. ** Une tonalité se fait entendre si la résistance mesurée est inférieure à environ 30 Ω. *** La tension de la diode est illustrée s’il y conductance, ou « OL » est affiché s’il n’y a pas conductance. † 3 200 µA max. pour le DM-100, 4 000 µA max. pour le DM-110. 58 COM DM-100 • DM-110 Utilisation (suite) Mise hors tension automatique Le multimètre s’éteint automatiquement après une certaine période d’inactivité. sur l’affichage indique que cette fonction est activée. Le multimètre émet un bip environ 15 secondes avant de s’éteindre. Le DM-100 s’éteint après environ 10 minutes et le DM-110 après 30 minutes. La lecture en cours est gardée en mémoire. Pour rallumer le multimètre, appuyer sur n’importe quel bouton ou tourner le sélecteur. La dernière lecture s’affiche en mode de maintien. Appuyer sur HOLD pour revenir au mode normal de fonctionnement. Mise hors tension automatique Pour désactiver la fonction de mise hors tension automatique, appuyer sur et maintenir le bouton RANGE en allumant le multimètre. Remarque : Lorsque le multimètre est mis hors tension puis rallumé de nouveau à l’aide du sélecteur, l’appareil revient au mode par défaut. Toutefois, la dernière valeur n’est pas gardée en mémoire. 59 Mesures types Mesure de la tension Mesure du courant 60 DM-100 • DM-110 Mesures types (suite) Mesure de la résistance Vérification de la continuité 61 Mesures types (suite) Mesure de la capacité Mesure des diodes Polarité Inversée Polarisation Directe 62 DM-100 • DM-110 Précision : DM-100 Pour obtenir les conditions d’utilisation et les coefficients de température, consulter la section sur les spécifications. La précision est spécifiée comme suit : ± (un pourcentage de la lecture + une quantité fixe) à 23 °C ± 5 °C (73,4 °F ± 9 ºF), 0 à 80 % d’humidité relative. Pour le DM-100, la précision du c.a. est spécifiée uniquement pour les ondes sinusoïdales. Courant c.a. Plage* Précision Tension de consommation 320,0 µA ± (2,2 % + 0,5 µA) 1 V max 3 200 µA ± (2,2 % + 5 µA) 10 V max 10,00 A ± (2,5 % + 0,05 A) 2 V max Tension c.a. Plage* Précision 3,200 V ± (1,50 % + 0,005 V) 32,00 V ± (1,50 % + 0,05 V) 320,0 V ± (1,50 % + 0,5 V) 750 V ± (1,50 % + 5 V) Plage de fréquence Impédance d’entrée 40 à 500 Hz 10 MΩ, 100 pF Courant c.c. Plage* Précision Tension de consommation 320,0 µA ± (1,7 % + 0,2 µA) 1 V max 3 200 µA ± (1,7 % + 2 µA) 10 V max 10,00 A ± (2 % + 0,02 A) 2 V max * En mode de sélection automatique de plage, les plages changent à environ tous les 3 200 comptes. En mode de sélection de plage manuel, la plage s’étend jusqu’à 3 400 comptes. 63 Précision : DM-100 (suite) Tension c.c. Plage* Précision Impédance d’entrée 320,0 mV ± (0,5 % + 0,2 mV) supérieur à 1 000 MΩ 3,200 V ± (0,5 % + 0,002 V) 32,00 V ± (0,5 % + 0,02 V) 320,0 V ± (0,5 % + 0,2 V) 1 000 V ± (0,5 % + 2 V) 10 MΩ Résistance Plage* Précision 320,0 Ω ± (1 % + 0,4 Ω) 3,200 kΩ ± (0,75 % + 0,003 kΩ) 32,00 kΩ ± (0,75 % + 0,03 kΩ) 320,0 kΩ ± (0,75 % + 0,3 kΩ) 3,200 MΩ ± (1 % + 0,003 MΩ) 32,00 MΩ ± (2 % + 0,05 MΩ) Tension typique de circuit ouvert : 1,5 V * En mode de sélection automatique de plage, les plages changent à environ tous les 3 200 comptes. En mode de sélection de plage manuel, la plage s’étend jusqu’à 3 400 comptes. Vérification de la diode Courant de vérification (Typique) : 1,5 mA Tension de circuit ouvert : Moins de 3,1 V c.c. Continuité Seuil : Une tonalité se fait entendre si la résistance mesurée est inférieure à environ 30 Ω. Temps de réponse : Environ 100 ms 64 DM-100 • DM-110 Précision : DM-110 Pour obtenir les conditions d’utilisation et les coefficients de température, consulter la section sur les spécifications. La précision est spécifiée comme suit : ± (un pourcentage de la lecture + une quantité fixe) à 23 °C ± 5 °C (73,4 °F ± 9 ºF), 0 à 80 % d’humidité relative. Pour le DM-110, la spécification de précision du c.a. s’applique aux ondes sinusoïdales à leur pleine échelle et aux ondes non-sinusoïdales inférieures à la demi-grandeur. • Pour les facteurs de crête entre 1,4 et 3, ajouter 1,5 % à la précision. • Pour les facteurs de crête entre 3 et 4, ajouter 3 % à la précision. Courant c.a. Plage Précision Tension de consommation 400,0 µA ± (2,2 % + 0,5 µA) 1 V max 4 000 µA ± (2,2 % + 5 µA) 10 V max 10,00 A* ± (2,5 % + 0,05 A) 2 V max * 10 A continu ; 20 A maximum (Durée de mise sous tension : 30 secondes sous tension, 5 minutes hors tension) Tension c.a. Plage Précision 400,0 mV non spécifié 4 000 V ± (1,5 % + 0,005 V) 40,00 V ± (1,3 % + 0,05 V) 400,0 V ± (1,3 % + 0,5 V) 750 V ± (1,3 % + 5 V) Plage de fréquence Impédance d’entrée 40 à 500 Hz 10 MΩ, 100 pF Courant c.c. Plage Précision 400,0 µA ± (1,7 % + 0,2 µA) Tension de consommation 1 V max 4 000 µA ± (1,7 % + 2 µA) 10 V max 10,00 A ± (2 % + 0,02 A) 2 V max 65 Précision : DM-110 (suite) Tension c.c. Plage Précision Impédance d’entrée 400,0 mV ± (0,5 % + 0,2 mV) supérieur 1 000 MΩ 4,000 V ± (0,5 % + 0,002 V) 40,00 V ± (0,5 % + 0,02 V) 400,0 V ± (0,5 % + 0,2 V) 1 000 V ± (0,5 % + 2 V) 10 MΩ Résistance Plage Précision 400,0 Ω ± (1,0 % + 0,3 Ω) 4,000 kΩ ± (0,75 % + 0,002 kΩ) 40,00 kΩ ± (0,75 % + 0,02 kΩ) 400,0 kΩ ± (0,75 % + 0,2 kΩ) 4,000 MΩ ± (1,0 % + 0,003 MΩ) 40,00 MΩ ± (1,5 % + 0,05 MΩ) Tension typique de circuit ouvert : 1,3 V Capacité Plage Précision 4,000 nF ± (3,0 % + 0,010 nF) 40,00 nF ± (3,0 % + 0,10 nF) 400,0 nF ± (2,0 % + 0,8 nF) 4,000 µF ± (2,0 % + 0,008 µF) 40,00 µF ± (2,0 % + 0,08 µF) 400,0 µF ± (2,0 % + 0,8 µF) 4,000 mF* ± (5,0 % + 0,020 mF) 40,00 mF* ± (5,0 % + 0,20 mF) Les précisions valent pour les condensateurs à couches (condensateurs dont l’absorption diélectrique est négligeable). *La précision spécifiée pour les valeurs inférieures à la moitié de la pleine échelle. Plusieurs peuvent fluctuer aux environs des limites de précision spécifiées. 66 DM-100 • DM-110 Précision : DM-110 (suite) Fréquence Plage Précision 4,000 kHz ± (0,01 % + 0,001 kHz) Sensibilité 40,00 kHz ± (0,01 % + 0,01 kHz) 400,0 kHz ± (0,01 % + 0,1 kHz) 4,000 MHz ± (0,01 % + 0,001 MHz) 250 mV rms 40,00 MHz ± (0,01 % + 0,01 MHz) 1 V rms 100 mV rms * *La sensibilité est de 1,5 V rms inférieure à 20 Hz La fonction de mesure de la fréquence comporte quatre sélections du niveau de la gâchette. Utiliser le bouton « Range » pour sélectionner. Vérification de la diode Courant de vérification (Typique) : 1,5 mA Tension de circuit ouvert : Moins de 3,1 V c.c. Continuité Seuil : Une tonalité se fait entendre si la résistance mesurée est inférieure à environ 30 Ω. Temps de réponse : Environ 100 ms 67 Spécifications Affichage : DM-100 : Affichage à cristaux liquides (3 200) et graphique à barres à 70 segments En mode de sélection automatique de plage, les plages changent à environ tous les 3 200 comptes. En mode de sélection de plage manuel, la plage s’étend jusqu’à 3 400 comptes. DM-110 : Affichage à cristaux liquides (4 000) et graphique à barres à 82 segments Polarité : Automatique Taux d’échantillonnage : Afficheur numérique : 2 par seconde Afficheur de graphique à barres : 12 par seconde Coefficient de température : 0,15 x (précision) par °C au-dessous de 18 °C ou au-dessus de 28 °C Mise hors tension automatique : DM-100 : après 10 minutes d’inactivité DM-110 : après 30 minutes d’inactivité Protections de surcharge : Volts : Consulter les Catégories de protection contre la surtension Ampères : A : Fusible F de 16 A/500 V HBC, pouvoir de coupure nominal 10 kA, 1/4 x 1-1/4 po Autres fonctions : 600 V c.c./V c.a. RMS Catégories de protection contre la surtension : Borne VΩµA : Catégorie III, 600 V c.a. et 600 V c.c. Catégorie II, 1 000 V c.a. et 1 000 V c.c. Conditions d’utilisation : 0 à 30 °C (32 à 86 °F), 0 à 80 % d’humidité relative 30 à 40 °C (86 à 104 °F), 0 à 75 % d’humidité relative 40 à 50 °C (104 à 122 °F), 0 à 45 % d’humidité relative Utilisation à l’intérieur uniquement. Conditions d’entreposage : -20 à 60 °C (-4 à 140 °F), 0 à 80 % d’humidité relative Enlever les piles Elévation : 2 000 m (6 500 pi) maximum Degré de pollution : 2 Pile : DM-100 : deux piles AAA de 1,5 volt (NEDA 24 A ou IEC LR03) DM-110 : une pile de 9 volts (NEDA 1604 ou IEC 6F22) Durée de vie de la pile : DM-100 : Environ 500 heures avec une pile alcaline DM-110 : Environ 300 heures avec une pile alcaline 68 DM-100 • DM-110 Catégories d’installation de surtension Ces définitions sont dérivées des normes internationales sur la sécurité pour la coordination de l’isolation telle qu’elle s’applique à la mesure, au contrôle et à l’équipement de laboratoire. Ces catégories de surtension sont expliquées plus en détail par Commission électrotechnique internationale ; se reporter à l’une de ces deux publications IEC 1010-1 ou IEC 60664. Surtension, catégorie I Niveau de signal. Pièces ou équipement électronique et de télécommunication. Par exemple, les circuits électroniques protégés contre les courants transitoires, dans les photocopieurs et les modems. Surtension, catégorie II Niveau local. Appareils, équipement portatif et les circuits dans lesquels ils sont branchés. Par exemple, les appareils d’éclairage, les téléviseurs et les dérivations. Surtension, catégorie III Niveau de distribution. Les machines installées en permanence et les circuits auxquels elles sont câblées. Par exemple, les systèmes de convoyeurs et les panneaux principaux de disjoncteurs du système électrique d’un édifice. Surtension, catégorie IV Niveau d’alimentation principal. Lignes surélevées et autres systèmes de câbles. Par exemple, les câbles, les multimètres, les transformateurs et autres équipements extérieurs appartenant aux fournisseurs en électricité. 69 Entretien • Ne pas tenter de réparer cet appareil. Il ne comporte aucune pièce pouvant être réparée. • Ne pas exposer l’appareil à des températures ou à une humidité extrêmes. Voir les spécifications. L’inobservation de ces consignes peut endommager l’appareil et entraîner des blessures. Remplacement de la pile et du fusible Avant d’ouvrir le boîtier, retirer les fils d’essai du circuit et mettre l’appareil hors tension. L’inobservation de cette consigne peut entraîner des blessures graves, voire mortelles. Le fusible est une partie intégrante du système de protection de surtension. Lorsqu’un fusible doit être remplacé, consulter les spécifications pour connaître le type, la taille et la capacité requis. L’utilisation de tout autre type de fusible annule le calibrage de protection contre la surtension de l’appareil. L’inobservation de cette consigne peut entraîner des blessures graves, voire mortelles. 1. Débrancher l’appareil du circuit. Mettre l’appareil hors tension (OFF). 2. Enlever les vis du couvercle arrière. 3. Retirer le couvercle arrière. 4. Lever le support de pile pour le sortir du couvercle arrière. 5. Remplacer les piles (selon la polarité) et/ou le fusible. 6. Replacer le couvercle, les vis et le support. Nettoyage Nettoyer régulièrement le boîtier avec un chiffon humide et un détergent doux. Ne pas utiliser de produits abrasifs ou de solvants. 70 DM-100 • DM-110 Greenlee Textron / Subsidiary of Textron Inc. 4455 Boeing Dr., Rockford, IL 61109-2988 815/397-7070 71 Lifetime Limited Warranty Greenlee warrants to the original purchaser of these goods for use that these products will be free from defects in workmanship and material for their useful life, excepting normal wear and abuse. This warranty is subject to the same terms and conditions contained in Greenlee’s standard one-year limited warranty. For all Test Instrument repairs, ship units Freight Prepaid to: Greenlee Textron, 4411 Boeing Drive, Rockford, IL 61109-2932 USA. Mark all packages: Attention TEST INSTRUMENT REPAIR. For items not covered under warranty (such as dropped, abused, etc.), repair cost quote available upon request. Note: Prior to returning any test instrument, please check replaceable batteries or make sure the battery is at full charge. Garantía limitada de por vida Greenlee garantiza al comprador original de estos productos para su uso que estos productos estarán libres de defectos de mano de obra y materiales durante toda su vida útil, exceptuando el desgaste normal y el abuso. Esta garantía está sujeta a los mismos términos y condiciones contenidos en la garantía estándar limitada de Greenlee de un año de duración. Para reparación de instrumentos de verificación, envíe las unidades con flete pagado a: Greenlee Textron, 4411 Boeing Drive, Rockford, IL 61109-2932 EE. UU. Marque todos los paquetes: Atención TEST INSTRUMENT REPAIR (Reparación de instrumentos de verificación). Para artículos no cubiertos por la garantía (tales como los que se han dejado caer o han sido maltratados, etc.) se puede cotizar el costo de la reparación a pedido. Nota: Antes de enviar cualquier instrumento de verificación, revise por favor las pilas o asegúrese de que estén totalmente cargadas. Garantie à vie limitée La société Greenlee garantit à l’acheteur d’origine de ces produits que ces derniers ne comportent aucun défaut d’exécution ou de matériau pour la durée de leur vie utile, sauf l’usure normale. Cette garantie est assujettie aux mêmes conditions que celles contenues dans les modalités et conditions de la garantie limitée standard d’un an de Greenlee. Pour toutes les réparations d’instruments de mesure, expédiez l’appareil en port payé à l’adresse suivante : Greenlee Textron, 4411 Boeing Drive, Rockford, IL 61109-2932, États-Unis. Sur tous les colis, inscrivez : Attention : TEST INSTRUMENT REPAIR (Réparation d'instrument de mesure). Lorsque les articles ne sont pas protégés par une garantie (comme si l’appareil est échappé, s’il est soumis à un usage abusif, etc.), une soumission pour le prix de réparation sera présentée sur demande. Remarque : Avant de renvoyer un appareil de mesure, vérifiez la pile et assurez-vous qu’elle est chargée au complet. For technical assistance: 800/435-0786 Greenlee Textron / Subsidiary of Textron Inc. 4455 Boeing Drive, Rockford, IL 61109-2988 USA Technical / Customer Service (International): 815/397-7070 • Fax: 815/397-9247 Customer Service (North America): 800/435-0786 • Fax: 800/451-2632, 815/397-1865 Canada Fax: 800/524-2853
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72

Greenlee DM-110 El manual del propietario

Categoría
Medir, probar
Tipo
El manual del propietario
Este manual también es adecuado para