Aim-TTI TGF4162 Guía de inicio rápido

Tipo
Guía de inicio rápido

Este manual también es adecuado para

FR
DE IT
ES
TGF4000 SERIES
40MHz, 80MHz, 160MHz & 240MHz
Dual Channel Arbitrary Funcon Generators
QUICK START GUIDE
EN
CONTENTS
Quick Guide in English ......................................................................................................................... 2
Guide Succinct en Français ................................................................................................................ 43
Kurzanleitung..................................................................................................................................... 83
Guida Rapida in Italiano ................................................................................................................... 123
G ................................................................................................................... 163
Specification .................................................................................................................................... 203
0 - Table of Contents
2 TGF4000 Series Quick Start Guide
TABLE OF CONTENTS
 Introduction ................................................................................................................................................. 3
The TGF4000 Series of Arbitrary Function Generators ...................................................................................... 3
About this Guide ............................................................................................................................................... 3
 Safety ........................................................................................................................................................... 4
General .............................................................................................................................................................. 4
Symbols ............................................................................................................................................................. 5
 Operational Principles .................................................................................................................................. 6
Front Panel Layout ............................................................................................................................................ 6
Rear Panel Layout .............................................................................................................................................. 7
Screen Layout .................................................................................................................................................... 8
 Getting Started ............................................................................................................................................. 9
 Basic Set-up Examples ................................................................................................................................ 10
Setting-up a Sine Wave Signal ......................................................................................................................... 10
Setting-up a Square Wave Clock Signal ........................................................................................................... 14
Setting-up a Pulse Waveform .......................................................................................................................... 20
Setting-up more Output Options .................................................................................................................... 28
Requirement ................................................................................................................................................... 28
 Exploring the Generator Capabilities .......................................................................................................... 32
Setting-up an arbitrary wave signal ................................................................................................................. 32
Setting-up an AM modulated Sine Waveform ................................................................................................. 33
PRBS ................................................................................................................................................................ 34
Frequency Modulation of a Sine Waveform .................................................................................................... 35
Pulse Width Modulated Waveform (PWM) ..................................................................................................... 36
Amplitude shift keying (ASK) ........................................................................................................................... 37
Frequency Sweep of a Sine Wave .................................................................................................................... 38
Generating a Triggered Burst........................................................................................................................... 39
Coupling the Frequency of Both Channels ...................................................................................................... 40
Frequency counter .......................................................................................................................................... 41
 Maintenance .............................................................................................................................................. 42
Cleaning .......................................................................................................................................................... 42
The latest revisions of this manual, device drivers and software tools can be downloaded from:
http://www.aimtti.com/support
1 - Introduction
The TGF4000 Series of Arbitrary Function Generators
3 TGF4000 Series Quick Start Guide
1. INTRODUCTION
The TGF4000 Series of Arbitrary Function Generators
This manual covers all four TGF4000 dual channel generators. Where there are differences in the
specification, the limits for the TGF4042 & TGF4082 are shown in square brackets [ ] after the
TGF4162 & TGF4242 limits.
These programmable function/arbitrary generators use direct digital synthesis techniques to
provide high performance and extensive facilities in a compact instrument. They generate a wide
variety of waveforms with high resolution and accuracy.
Sine waves are produced with low distortion to 160MHz/240MHz [40MHz/80MHz]. Square waves
have fast rise and fall times at up to 100MHz [25MHz]. Linear ramp waves are produced to 5MHz.
Ramp and square waves also have variable symmetry.
The instruments generate high resolution, low jitter, variable edge time pulses to 100MHz [25MHz]
with variable period, pulse width, pulse delay, pulse edges and amplitude. Complex custom
waveforms can be generated with 16-bit [14-bit] resolution and a sampling rate of 800MSa/s
[400MSa/s]. Up to four waveforms can be stored in internal memory. Waveforms can also be
generated by the supplied Waveform Manager Plus V4.13 Windows application and downloaded to
the instrument via USB, LAN or optional GPIB interfaces or via a USB flash drive.
Front panel operation is straightforward and user friendly with all major parameters shown at all
times on the large, bright, colour LCD. All major functions can be accessed with a single key or two.
The knob or numeric keypad can be used to adjust frequency, amplitude, offset, and other
parameters. Voltage values can be entered directly in Vpp or as high and low levels. Timing
parameters can be entered in Hertz (Hz) or seconds.
Internal AM, FM, PM, ASK, FSK, BPSK, SUM* and PWM modulation make it easy to modulate
waveforms without the need for a separate modulation source. Linear and logarithmic sweeps are
also built in, with sweep rates selectable from 1 µs to 500s. Burst mode operation allows for a user-
selected number of cycles at each trigger event.
LAN and USB interfaces are standard and there is full compliance to 1.5 LXI Device Specification
2016.
The instruments use a high stability temperature compensated internal oscillator and the external
frequency reference input lets you synchronize to an external 10 MHz frequency standard for even
greater accuracy.
*TGF4162 & TGF4242 only
About this Guide
This Quick Start guide is for bench-top use of the TGF4000 Series comprising the
TGF4042,TGF4082, TGF4162 and TGF4242 dual channel generators. A full Instruction Manual
(English only) is also provided on the Aim-TTi website that includes comprehensive explanations of
all functions and additional information on remote control, calibration, and the detailed technical
specifications.
2 - Safety
General
4 TGF4000 Series Quick Start Guide
2. SAFETY
General
This generator is a Safety Class I instrument according to IEC classification and has been designed to
meet the requirements of EN610101 (Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use). It is an Installation Category II instrument intended for
operation from a normal single phase supply.
This instrument has been tested in accordance with EN610101 and has been supplied in a safe
condition. This instruction manual contains some information and warnings which have to be
followed by the user to ensure safe operation and to retain the instrument in a safe condition.
This instrument has been designed for indoor use in a Pollution Degree 2 environment in the
temperature range 5°C to 40°C, 20% 80% RH (noncondensing). It may occasionally be subjected
to temperatures between +5° and 10°C without degradation of its safety. Do not operate while
condensation is present.
Use of this instrument in a manner not specified by these instructions may impair the safety
protection provided. Do not operate the instrument outside its rated supply voltages or
environmental range.
WARNING! THIS INSTRUMENT MUST BE EARTHED
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited. The protective action must not be
negated by the use of an extension cord without a protective conductor.
When the instrument is connected to its supply, terminals may be live and opening the covers or
removal of parts (except those to which access can be gained by hand) is likely to expose live parts.
The apparatus shall be disconnected from all voltage sources before it is opened for any
adjustment, replacement, maintenance or repair. Any adjustment, maintenance and repair of the
opened instrument under voltage shall be avoided as far as possible and, if inevitable, shall be
carried out only by a skilled person who is aware of the hazard involved.
If the instrument is clearly defective, has been subject to mechanical damage, excessive moisture
or chemical corrosion the safety protection may be impaired and the apparatus should be
withdrawn from use and returned for checking and repair.
Make sure that only fuses with the required rated current and of the specified type are used for
replacement. The use of makeshift fuses and the shortcircuiting of fuse holders is prohibited.
This instrument uses a Lithium button cell for nonvolatile memory battery backup; typical life is
5 years. In the event of replacement becoming necessary, replace only with a cell of the correct
type, i.e. 3V Li/Mn0
2
20mm button cell type 2032. Exhausted cells must be disposed of carefully in
accordance with local regulations; do not cut open, incinerate, expose to temperatures above 60°C
or attempt to recharge.
Do not wet the instrument when cleaning it and in particular use only a soft dry cloth to clean the
LCD window.
2 - Safety
Symbols
5 TGF4000 Series Quick Start Guide
Symbols
The following symbols are used on the instrument and in this manual:
Caution
refer to the accompanying documentation,
incorrect operation may damage the instrument.
Terminal connected to chassis ground.
Mains supply OFF.
l
Mains supply ON.
Alternating current.
3 - Operational Principles
Front Panel Layout
6 TGF4000 Series Quick Start Guide
3. OPERATIONAL PRINCIPLES
Front Panel Layout
Ref.
Short Description
Function
1
Power Switch
Switches instrument on or off. Safety Note: To fully disconnect from
the AC supply, unplug the mains cord from the back of the
instrument or switch off at the AC supply outlet; make sure that the
means of disconnection is readily accessible.
2
Soft-keys
Performs the function shown on the LCD soft-key label above.
3
Waveform Keys
Selects the main waveform type (carrier waveform) as active. (Sine,
Square, Ramp, Pulse, Noise/PBRS or Arb.)
4
Waveform Modification
Menus
Opens menus for setting parameters for Modulation, Sweep and
Burst
5
Other Menus
Selects menus for internal and external file storage, instrument
utilities, and trigger conditions.
6
Main Sockets
Main output sockets. Channel 2 can also be configured to output
Channel 1 sync from its MAIN OUT 2 socket.
7
Output Keys
Switch the selected MAIN OUT on or off.
8
Cursor Keys and Spin Wheel
Used to change numeric parameter values digit by digit.
Used to select items within some menus.
9
Numeric Keypad
Used to enter numeric parameter values directly.
10
USB Flash Drive
USB Host connector for USB Flash drive storage.
3 - Operational Principles
Rear Panel Layout
7 TGF4000 Series Quick Start Guide
Rear Panel Layout
Ref.
Short Description
1
Modulation Input
2
Reference In / AC coupled
frequency counter
frequency measurement.
3
Reference Out
4
Trigger Input / DC coupled
frequency counter
external frequency measurement
5
LAN connection
the TCP/IP Socket protocol.
6
USB connection
7
GPIB connection (optional)
3 - Operational Principles
Screen Layout
8 TGF4000 Series Quick Start Guide
Screen Layout
Ref.
Short Description
Function
1
Channel Indicator
Shows currently selected channel
2
Main Waveform type
Shows current carrier waveform
3
Output State
Shows main output On or Off
4
External Clock Indicator
Shows status of external clock (if applied)
5
LAN Status Indicator
Shows status of LAN (Ethernet) connection.
6
Parameters Box
Shows main parameters for waveform.
7
Menu Description
Shows the currently selected editing menu.
8
Graph Box
Shows a graphical representation of the selected waveform.
9
Edit Box
Shows the current parameter that can be edited
10
Soft-key Labels
Shows the current functions for the six keys below.
4 - Getting Started
Initial Conditions
9 TGF4000 Series Quick Start Guide
4. GETTING STARTED
In order to familiarise the user with some of the basic functionalities of the instrument, a number
of set-up examples are shown in this guide.
It is recommended that all users should carry out the first four examples:
Setting-up a Sine Wave Signal
S
etting-up a Square Wave Clock Signal
S
etting-up a Pulse Waveform
S
etting-up more Output Options
A number of further set-up examples are provided that assume some familiarity with the
instrument:
Setting-up an arbitrary wave signal
Setting-up an AM modulated Sine Waveform
P
RBS (TGF4162 & TGF4242)
Frequency Modulation of a Sine Waveform
Pulse Width Modulated Waveform (PWM)
Ampl
itude shift keying (ASK)
F
requency Sweep of a Sine Wave
G
enerating a Triggered Burst
C
oupling the Frequency of Both Channels
F
requency counter
F
or more detailed information on all functionality - see the full Instruction Manual.
Initial Conditions
Before setting up the instrument for any of the examples, it should be returned to default
conditions. To do this follow these steps:
Press the hard key marked Utility
Press the soft-key labelled System
Press the soft-key labelled Default (display will show Restore Factory Default?)
Press the soft-key labelled Yes
This sets the main waveform to Sine (10kHz, 1V pk-pk) and cancels any modulations, sweep, or
burst triggering or gating.
NOTE
The instrument can be set to remember its latest settings on power-off and restore them at power-
on. This is set from the Utility > System menu and the PwrOn soft-key. This setting will be lost
when the instrument is restored to default conditions as described above.
5 - Basic Set-up Examples
Setting-up a Sine Wave Signal
10 TGF4000 Series Quick Start Guide
5. BASIC SET-UP EXAMPLES
Setting-up a Sine Wave Signal
Requirement
Output a continuous sine wave signal with 40MHz frequency and an amplitude of 6 volts pk-pk
from MAIN OUT 1.
Starting Conditions
Before starting, reset the instrument to defaults as described in section 4 Getting Started
Open Waveform Menu - Sine
Press the hard key marked Waves
Press the soft key labelled Sine
5 - Basic Set-up Examples
Setting-up a Sine Wave Signal
11 TGF4000 Series Quick Start Guide
Set the Frequency
The soft key labelled Freq will be highlighted - the current frequency appears in the edit box.
Pressing this soft-key repeatedly changes its function between Frequency and Period.
Use the numeric keypad to enter a new frequency. Press the numbers 4 0
As soon as a number is entered, the soft-keys change to show units of frequency.
Press the soft-key labelled MHz to confirm a frequency of 40MHz.
5 - Basic Set-up Examples
Setting-up a Sine Wave Signal
12 TGF4000 Series Quick Start Guide
Set the Amplitude
Press the soft key labelled Ampl
Successive presses of the Ampl soft-key changes the Ampl and Offset key labels to
HiLvl (high level) and LoLvl (low level) and vice versa.
Use the numeric keypad to enter a new amplitude. Press the number 6
As soon as a number is entered, the soft-keys change to show units of voltage.
5 - Basic Set-up Examples
Setting-up a Sine Wave Signal
13 TGF4000 Series Quick Start Guide
Press the soft-key labelled Vpp to confirm a pk-pk amplitude of 6.0 volts.
Turn the Output On
Press Output 1 key to turn the channel 1 output On.
The Output 1 key illuminates orange to indicate the on state.
5 - Basic Set-up Examples
Setting-up a Square Wave Clock Signal
14 TGF4000 Series Quick Start Guide
Setting-up a Square Wave Clock Signal
Requirement
Output a continuous square wave clock signal with 20MHz frequency, 50% duty cycle and a high
level of 3.3V and a low level of 0.0 volts from MAIN OUT 1.
Starting Conditions
Before starting, reset the instrument to defaults as described in section 4 Getting Started
Open Waveform Menu - Square
Press the hard key marked Waves
Press the soft-key labelled Square.
+
5 - Basic Set-up Examples
Setting-up a Square Wave Clock Signal
15 TGF4000 Series Quick Start Guide
Set the Frequency
The soft key labelled Freq will be highlighted - the current frequency appears in the edit box.
Pressing this soft-key repeatedly changes its function between Frequency and Period.
Use the numeric keypad to enter a new frequency. Press the numbers 2 0
As soon as a number is entered, the soft-keys change to show units of frequency.
Press the soft-key labelled MHz to confirm a frequency of 20MHz.
The graph box changes to show the rise time on the edges which is now significant.
5 - Basic Set-up Examples
Setting-up a Square Wave Clock Signal
16 TGF4000 Series Quick Start Guide
Confirm the Duty Cycle
Press the soft-key labelled Duty - the current duty cycle appears in the edit box.
The duty cycle is already set at 50%, but could be changed here if required.
Set the High and Low Levels
Press the soft-key labelled Ampl - the key label changes to HiLvl and the current high level
voltage appears in the edit box.
Successive presses of the Ampl soft-key changes the Ampl and Offset key labels to
HiLvl (high level) and LoLvl (low level) and vice versa.
5 - Basic Set-up Examples
Setting-up a Square Wave Clock Signal
17 TGF4000 Series Quick Start Guide
When the soft-key is labelled HiLvl - the current high level voltage appears in the edit box.
Use the numeric keypad to enter a new level. Press the numbers 3 . 3
As soon as a number is entered, the soft-keys change to show units of voltage.
Press the soft-key labelled V to confirm a high level of 3.3 volts.
5 - Basic Set-up Examples
Setting-up a Square Wave Clock Signal
18 TGF4000 Series Quick Start Guide
Press the soft-key labelled LoLvl - the current low level voltage appears in the edit box.
Use the numeric keypad to enter a new level. Press 0
Press the soft-key labelled V to confirm a low level of 0.0 volts.
Turn the Output On
Press Output 1 key to turn the channel 1 output On.
The Output 1 key illuminates orange to indicate the on state
5 - Basic Set-up Examples
Setting-up a Square Wave Clock Signal
19 TGF4000 Series Quick Start Guide
Making live changes to any numeric parameter (e.g. Frequency)
Numeric parameters can be changed by using the cursor keys and spin wheel as an alternative to
the numeric keypad.
Press the hard key marked Waves
Press the soft-key labelled Square.
Press the soft-key labelled Freq the current frequency value of 20.0MHz is displayed
Press the Cursor hard keys to move the edit highlight to the second digit.
Use the spin wheel to change the value the frequency is changed immediately.
NOTE
Press to activate/ deactivate the spin wheel.
5 - Basic Set-up Examples
Setting-up a Pulse Waveform
20 TGF4000 Series Quick Start Guide
Setting-up a Pulse Waveform
Requirement
Output a continuous pulse signal with 100ns period, 30ns pulse width, 20ns edge times and a high
level of 2.7V and a low level of -0.6 volts from MAIN OUT 1.
Starting Conditions
Before starting, reset the instrument to defaults as described in section 4 Getting Started
Open Waveform Menu - Pulse
Press the hard key marked Waves
Press the soft-key labelled Pulse.
5 - Basic Set-up Examples
Setting-up a Pulse Waveform
21 TGF4000 Series Quick Start Guide
Set the Period
Press the soft-key labelled PlsFrq so that it changes to PlsPer - the current period appears
in the edit box.
Pressing this soft-key repeatedly changes its function between Frequency and Period.
Use the numeric keypad to enter a new period. Press the numbers 1 0 0
As soon as a number is entered, the soft-keys change to show units of time.
5 - Basic Set-up Examples
Setting-up a Pulse Waveform
22 TGF4000 Series Quick Start Guide
Press the soft-key labelled ns to confirm a period of 100ns.
The graph box changes to show a representation of the pulse and edge times.
Set the Pulse Width
Press the soft-key labelled Duty - the key label changes to Width and displays the width as
a time
5 - Basic Set-up Examples
Setting-up a Pulse Waveform
23 TGF4000 Series Quick Start Guide
Use the numeric keypad to enter a new width. Press the numbers 3 0.
As soon as a number is entered, the soft-keys change to show units of time.
Press the soft-key labelled ns to confirm a width of 30ns.
5 - Basic Set-up Examples
Setting-up a Pulse Waveform
24 TGF4000 Series Quick Start Guide
Set the Pulse Edge Times
Press the soft-key labelled More
Press the soft-key labelled Edge
Use the cursor keys to select the digit representing units of 10ns
Use the spin wheel to change the value to 20.0ns
The value could have been entered using the numeric keypad if preferred.
Press the soft-key labelled Done
5 - Basic Set-up Examples
Setting-up a Pulse Waveform
25 TGF4000 Series Quick Start Guide
Set the High and Low Levels
Press the soft-key labelled Ampl - the key label changes to HiLvl and the current high level
voltage appears in the edit box
Successive presses of the Ampl soft-key changes the Ampl and Offset key labels to
HiLvl (high level) and LoLvl (low level) and vice versa.
Use the numeric keypad to enter a new level. Press the numbers 2 . 7 .
As soon as a number is entered, the soft-keys change to show units of voltage.
5 - Basic Set-up Examples
Setting-up a Pulse Waveform
26 TGF4000 Series Quick Start Guide
Press the soft-key labelled V to confirm a high level of 2.7 volts.
Press the soft-key labelled LoLvl - the current low level voltage appears in the edit box.
5 - Basic Set-up Examples
Setting-up a Pulse Waveform
27 TGF4000 Series Quick Start Guide
Use the numeric keypad to enter a new level. Press - . 6
Press the soft-key labelled V to confirm a low level of -600 mV.
Turn the Output On
Press Output1 key to turn the channel 1 output On.
The Output 1 key illuminates orange to indicate the on state
5 - Basic Set-up Examples
Setting-up more Output Options
28 TGF4000 Series Quick Start Guide
Setting-up more Output Options
Requirement
In the earlier set-up examples it was shown how the output menu is used to set the output level
(amplitude plus offset or high level plus low level) and turn the output on or off. This example
demonstrates the setting of output phase, output polarity, load impedance and voltage auto-range.
Starting Conditions
Before starting, reset the instrument to defaults as described in section 4 Getting Started
Open the Output Menu
Press the soft-key marked Output
5 - Basic Set-up Examples
Requirement
29 TGF4000 Series Quick Start Guide
Change the Output Phase
The soft-key labelled Phase will be selected as default.
Enter a phase of -45 degree.
The set phase angle is the point in the waveform period which is coincident with the Sync or trigger
edge, i.e. it is the point in the period at which the waveform starts. Hence a negative phase setting
advances, and a positive phase setting delays the waveform relative to the Sync or trigger; the
waveform in the graph box changes to show this.
The Align soft-key is used to re-align phase when making frequency changes.
Change the Output Polarity
Press the soft-key labelled Type to invert the output polarity.
Successive presses of the type key alternates between normal and inverted.
5 - Basic Set-up Examples
Requirement
30 TGF4000 Series Quick Start Guide
Change the Load Impedance
Press the soft-key labelled Load
Press the soft-key labelled Load again to change the load impedance to High-z (high
impedance).
The default load impedance is 50 Ohms, but that this could be changed to any impedance between
50 and 10,000 Ohms. Levels are calculated based upon this impedance.
Successive presses of the Load key alternates between a numeric value and High-z.
Note that the amplitude readout increases to 2 volts pk-pk.
Press the soft-key labelled Load to return the load impedance to 50 Ohms.
5 - Basic Set-up Examples
Requirement
31 TGF4000 Series Quick Start Guide
Changing the Range
Press the soft-key labelled Range
Press the soft-key labelled Range again to change the range from Auto to Hold.
Auto mode auto-ranges in 6dB attenuator steps (i.e. rangemaximums of 10Vpp, 5Vpp, 2.5Vpp,
etc., into with the amplitude range limited to 6dB to maintain waveform quality.
Selecting Hold mode disables auto-ranging; the attenuator setting is fixed and the amplitude
range is no longer limited.
With range set to Auto the amplitude and attenuators will switch automatically and optimal
performance will be realised. With range set to Hold a fixed attenuator setting is used for all
amplitude settings.
6 - Exploring the Generator Capabilities
Setting-up an arbitrary wave signal
32 TGF4000 Series Quick Start Guide
6. EXPLORING THE GENERATOR CAPABILITIES
In the following examples only the parameter settings are described, together with the related key
names. The resultant output waveforms are shown, along with the sync or trigger waveform where
relevant. Output amplitude and offset settings are examples only and need not be followed.
Setting-up an arbitrary wave signal
Start with the instrument returned to Default Settings.
MENU
HARD KEY NAME
Waveform
Waves
MENU
Soft-key Name
Arbitrary waveform
Arb
Parameter
Soft-key Name
Setting
Wave selection
Waves
Sinc
Frequency
-
10kHz
Amplitude
Ampl
2V
Offset
Offset
5mVdc
Parameter
HARD KEY NAME
Setting
Output State
Output1
On
6 - Exploring the Generator Capabilities
Setting-up an AM modulated Sine Waveform
33 TGF4000 Series Quick Start Guide
Setting-up an AM modulated Sine Waveform
Start with the instrument returned to Default Settings. .
Parameter
Soft-key Name
Setting
Frequency
-
10MHz
MENU
HARD KEY NAME
Modulation
Mod
Parameter
Soft-key Name
Setting
Frequency
-
100kHz
Depth
Depth
100%
Source
Source
Internal
Shape
Shape
Sine
Modulation state
On/Off
On
Parameter
HARD KEY NAME
Setting
Output State
Output1
On
Change the modulation to AM-SC
Parameter
Soft-key Name
Setting
Type
Type > AM
AM-SC
6 - Exploring the Generator Capabilities
PRBS
34 TGF4000 Series Quick Start Guide
PRBS
Start with the instrument returned to Default Settings.
MENU
HARD KEY NAME
Waveform
Waves
Menu
Soft-key name
Setting
Noise/PRBS
Noise
Source
Source
PBRS
Parameter
Soft-key Name
Setting
Bit Rate
BitRate
1Mbps
Amplitude
Ampl
3.3V
Offset
Offset
1.65V
PRBS Type
Type
PN7
Parameter
HARD KEY NAME
Setting
Output State
Output1
On
6 - Exploring the Generator Capabilities
Frequency Modulation of a Sine Waveform
35 TGF4000 Series Quick Start Guide
Frequency Modulation of a Sine Waveform
Start with the instrument returned to Default Settings.
MENU
HARD KEY NAME
Modulation
Mod
Parameter
Soft-key Name
Setting
Modulation State
On/Off
On
Modulation Type
Type
FM
Modulation Frequency
-
1kHz
Deviation
Deviatn
9kHz
MENU
HARD KEY NAME
Sine
Params
Parameter
Soft-key Name
Setting
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
HARD KEY NAME
Setting
Output State
Output1
On
6 - Exploring the Generator Capabilities
Pulse Width Modulated Waveform (PWM)
36 TGF4000 Series Quick Start Guide
Pulse Width Modulated Waveform (PWM)
Start with the instrument returned to Default Settings.
MENU
HARD KEY NAME
Waveform
Waves
Parameter
Soft-key Name
Setting
Pulse
Pulse
MENU
HARD KEY NAME
Modulation
Mod
Parameter
Soft-key Name
Setting
Modulation State
On/Off
On
Modulation Type
Type
PWM
Modulation Frequency
-
1kHz
Deviation
Dev %
40%
MENU
HARD KEY NAME
Pulse
Params
Parameter
Soft-key Name
Setting
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
HARD KEY NAME
Setting
Output State
Output1
On
6 - Exploring the Generator Capabilities
Amplitude shift keying (ASK)
37 TGF4000 Series Quick Start Guide
Amplitude shift keying (ASK)
Start with the instrument returned to Default Settings.
MENU
HARD KEY NAME
Modulation
Mod
Parameter
Soft-key Name
Setting
Modulation State
On/Off
On
Modulation type
Type
ASK
Modulation Source
Source
Internal
Hop Amplitude
HpAmpl
100mV
Switching Rate
Rate
1kHz
Hop Polarity
HopPol
Positive
MENU
HARD KEY NAME
Sine
Params
Parameter
Soft-key Name
Setting
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
HARD KEY NAME
Setting
Output State
Output1
On
Note that the second trace is the output from the main Output 2 socket when Channel 2 is
synchronised which follows the modulating waveform signal.
6 - Exploring the Generator Capabilities
Frequency Sweep of a Sine Wave
38 TGF4000 Series Quick Start Guide
Frequency Sweep of a Sine Wave
Start with the instrument returned to Default Settings.
MENU
HARD KEY NAME
Sweep
Sweep
Parameter
Soft-key Name
Setting
Sweep State
On/Off
On
Stop Frequency
Freq > Stop
100kHz
MENU
HARD KEY NAME
Sine
Params
Parameter
Soft-key Name
Setting
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
HARD KEY NAME
Setting
Output State
Output1
On
6 - Exploring the Generator Capabilities
Generating a Triggered Burst
39 TGF4000 Series Quick Start Guide
Generating a Triggered Burst
Start with the instrument returned to Default Settings.
Parameter
Soft-key Name
Setting
Frequency
-
6MHz
MENU
HARD KEY NAME
Burst
Burst
Parameter
Soft-key Name
Setting
Burst State
On/Off
On
Burst Count
Count
3
Trigger Source
SetTrg > Source > Int
Internal Trigger
Trigger Period
SetTrg > Period
5ms
MENU
HARD KEY NAME
Sine
Params
Parameter
Soft-key Name
Setting
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
HARD KEY NAME
Setting
Output State
Output1
On
Note that the second trace is the output from the Main Output 2 socket when Channel 2 is
synchronised which follows the trigger input signal.
6 - Exploring the Generator Capabilities
Coupling the Frequency of Both Channels
40 TGF4000 Series Quick Start Guide
Coupling the Frequency of Both Channels
Start with the instrument returned to Default Settings.
MENU
HARD KEY NAME
Utility
Utility
Parameter
Soft-key Name
Setting
Frequencies
Dual Ch > Freq
Coupled
Parameter
HARD KEY NAME
Setting
Output State
Output 1
On
Output State
Output 2
On
MENU
HARD KEY NAME
Sine
Params
Parameter
Soft-key Name
Setting
Phase Shift
Output > Phase > Done
90 degrees
Frequency
Freq
1MHz
Note that, when channel 1 is set to 1MHz, channel 2 is also set to 1MHz.
The 90 degree phase shift between the channels can be seen.
6 - Exploring the Generator Capabilities
Frequency counter
41 TGF4000 Series Quick Start Guide
Frequency counter
Start with the instrument returned to Default Settings.
MENU
HARD KEY NAME
Utility
Utility
Parameter
Soft-key Name
Setting
Counter
Instr > FrCntr > On/Off
Counter enabled
Source
Source
TRIG IN - DC coupled
Type
Freq
Frequency
Measurement
Count
The Edit Box shows the current measurement. With no input signal at the selected input, the
counter displays No signaluntil an adequate input signal is applied. When an adequate input
signal is applied, the counter constantly measures and displays the reading in the Edit Box.
7 - Maintenance
Cleaning
42 TGF4000 Series Quick Start Guide
7. MAINTENANCE
The Manufacturers or their agents overseas will provide a repair service for any unit developing a
fault. Where owners wish to undertake their own maintenance work, this should only be done by
skilled personnel in conjunction with the service guide which may be obtained directly from the
Manufacturers or their agents overseas.
Cleaning
If the instrument requires cleaning use a cloth that is only lightly dampened with water or a mild
detergent.
WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE INSTRUMENT, NEVER ALLOW WATER TO
GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE NEVER CLEAN WITH SOLVENTS.
0 - Table des matières
43 TGF4000 Series Quick Start Guide
TABLE DES MATIÈRES
 Sécuri ............................................................................................................................................................ 44
Symboles .............................................................................................................................................................. 45
 Principes de fonctionnement ........................................................................................................................... 46
Disposition du panneau avant .............................................................................................................................. 46
Disposition du panneau arrière ............................................................................................................................ 47
Disposition de l'écran ........................................................................................................................................... 48
 Première utilisation .......................................................................................................................................... 49
Conditions initiales ............................................................................................................................................... 49
 Exemples de configuration de base .................................................................................................................. 50
Configurer un signal d'onde sinusoïdale ............................................................................................................... 50
Configurer un signal d'horloge à onde carrée ...................................................................................................... 54
Configuration d'une forme d'onde à impulsion .................................................................................................... 60
Configurer de plus amples options pour les sorties ............................................................................................. 68
 Explorer les fonctionnalités du générateur ....................................................................................................... 72
Configurer un signal d'onde arbitraire .................................................................................................................. 72
Configurer une forme d'onde sinusoïdale modulée AM ...................................................................................... 73
PRBS ..................................................................................................................................................................... 74
Modulation de fréquence d'une forme d'onde sinusoïdale ................................................................................. 74
Forme d'onde modulée à largeur d'impulsion (PWM) ......................................................................................... 76
Modulation par déplacement damplitude (FSK) ................................................................................................. 77
Balayage de fréquence d'une onde sinusoïdale ................................................................................................... 78
Génération d'une rafale déclenchée .................................................................................................................... 78
Couplage des fréquences de deux voies .............................................................................................................. 80
Compteur de fréquence ....................................................................................................................................... 81
 Maintenance .................................................................................................................................................... 82
Nettoyage ............................................................................................................................................................. 82
Les dernières versions de ce manuel, les pilotes de périphérique et les outils logiciels peuvent être
téléchargés à partir de: www.aimtti.com
1 - Sécurité
44 TGF4000 Series Quick Start Guide
1. SÉCURITÉ
Cet instrument est de Classe de sécurité 1 suivant la classification IEC et il a été construit pour
satisfaire aux impératifs EN61010-1 (Impératifs de sécurité pour le matériel électrique en vue de
mesure, commande et utilisation en laboratoire). Il s
agit dun instrument dinstallation Catégorie II
devant être exploité depuis une alimentation monophasée habituelle.
Cet instrument a é soumis à des essais conformément à EN61010-1 et il a été fourni en tout état
de sécurité. Ce manuel d
instructions contient des informations et avertissements qui doivent être
suivis par l
utilisateur afin dassurer un fonctionnement en toute sécurité et de conserver
l
instrument dans un état de bonne sécurité.
Cet instrument a été conçu pour être utilisé en interne dans un environnement de pollution
Degré 2, plage de températures 5°C à 40°C, 20% - 80% HR (sans condensation). Il peut être soumis
de temps à autre à des températures comprises entre +5°C et 10°C sans dégradation de sa
sécurité. Ne pas l
utiliser lorsquil y a de la condensation.
Toute utilisation de cet instrument de manière non spécifiée par ces instructions risque d
affecter
la protection de sécurité conférée. Ne pas utiliser l
instrument à lextérieur des tensions
d
alimentation nominales ou de la gamme des conditions ambiantes spécifiées.
AVERTISSEMENT! CET INSTRUMENT DOIT ETRE RELIE A LA TERRE
Toute interruption du conducteur de terre secteur à lintérieur ou à lextérieur de linstrument
rendra l
instrument dangereux. Il est absolument interdit deffectuer une interruption à dessein.
Ne pas utiliser de cordon de prolongation sans conducteur de protection, car ceci annulerait sa
capacité de protection.
Lorsque l
instrument est relié à son alimentation, il est possible que les bornes soient sous tension
et par suite, l
ouverture des couvercles ou la dépose de pièces (à lexception de celles auxquelles
on peut accéder manuellement) risque de mettre à découvert des pièces sous tension. Il faut
débrancher toute source de tension éventuelle de l
appareil avant de louvrir pour effectuer des
réglages, remplacements, travaux d
entretien ou deparation.
Eviter dans la mesure du possible d
effectuer des réglages, travaux de réparation ou dentretien
lorsque l
instrument ouvert est branché à une source dalimentation, mais si cest absolument
nécessaire, seul un technicien compétent au courant des risques encourus doit effectuer ce genre
de travaux.
S
il est évident que linstrument est défectueux, quil a été soumis à des dégâts mécaniques, à une
humidité excessive ou à une corrosion chimique, la protection decurité sera amoindrie et il faut
retirer l
appareil, afin quil ne soit pas utilisé, et le renvoyer en vue de vérifications et de
réparations.
Uniquement remplacer les fusibles par des fusibles d
intensité nominale requise et de type
spécifié. Il est interdit d
utiliser des fusibles bricolés et de court-circuiter des porte-fusibles.
L
instrument utilise une pile bouton au lithium pour la mémoire non-volatile ; sa durée de vie est
environ 5 ans. Pour son remplacement, utilisé une pile du même type : 3V Li/ Mn0
2
type 2032. Les
piles usées doivent être jetées en accord avec les lois locales ; ne pas les couper, les brûler, les
exposer à des températures au delà de 60°C ou essayer de la recharger.
Ne pas mouiller l
instrument lors de son nettoyage; en particulier, nutiliser quun chiffon doux et
sec pour nettoyer la vitre de l
afficheur.
1 - Sécurité
Symboles
45 TGF4000 Series Quick Start Guide
Symboles
Les symboles suivants se trouvent sur l
instrument, ainsi que dans ce manuel.
ATTENTION
- se référer à la documentation ci-jointe; toute
utilisation incorrecte risque d
endommager lappareil.
Borne reliée à la terre du châssis
Alimentation secteur ON (allumée)
l
Alimentation secteur OFF (éteinte)
Courant alternatif (c.a.)
2 - Principes de fonctionnement
Disposition du panneau avant
46 TGF4000 Series Quick Start Guide
2. PRINCIPES DE FONCTIONNEMENT
Disposition du panneau avant
Réf.
Brève description
Fonction
1
Commutateur
d'alimentation
Allume et éteint l'appareil. Remarque de sécurité : Afin de déconnecter
complètement lappareil de lalimentation CA, débrancher le cordon
d
alimentation à larrière de lappareil ou éteindre la prise dalimentation
CA ; il convient de s
assurer que l'élément utilisé pour déconnecter
l'appareil est accessible.
2
Touches programmables
Exécutent la fonction indiquée sur l'étiquette située au-dessus de la
touche programmable sur l'écran LCD.
3
Sélection des formes
d'ondes
Ouvre les menus de formes d'ondes pour régler les paramètres des
formes d'ondes.(sinusoïdales, carrées, rampes, impulsions, Bruit/PBRS
ou arb.)
4
Menus de modification
des formes d'onde
Ouvre les menus pour définir les paramètres de modulation, de balayage
et de rafales
5
Autres menus
Sélectionne les menus relatifs aux espaces de stockage internes et
externes, aux utilitaires de l'appareil et aux conditions de
déclenchement.
6
Prises principales
Prises de sortie principales La voie 2 peut également être configurée
pour produire la synchronisation de la voie 1 à partir de sa prise MAIN
OUT 2 (SORTIE PRINCIPALE 2).
7
Touches de sortie
Active ou désactive la sortie principale MAIN OUT sélectionnée.
8
Touches du curseur et
molette
Utilisées pour changer les valeurs des paramètres numériques chiffre
par chiffre. Utilisée pour sélectionner des éléments dans certains menus.
9
Pavé numérique
Utilisé pour saisir les valeurs de paramètres numériques directement.
10
Clé USB
Connecteur hôte USB pour disque de stockage USB.
2 - Principes de fonctionnement
Disposition du panneau arrière
47 TGF4000 Series Quick Start Guide
Disposition du panneau arrière
Réf.
Brève description
Fonction
1
Entrée de modulation
Entrée externe pour la modulation des formes d'ondes
principales.
2
Entrée de référence / Compteur
de fréquence de couplage CA
Entrée pour lhorloge de référence externe de 10 MHz et la
mesure de la fréquence externe de couplage CA.
3
Sortie de référence
Entrée pour une horloge de férence interne de 10 MHz.
4
Entrée de déclenchement /
Compteur de fréquence de
couplage CC
Entrée externe de déclenchement des formes dondes
principales et de mesure de la fréquence externe de couplage CC
5
LAN
L’interface LAN (réseau local) est conçue pour répondre aux
critères LXI Core 2011. Il est possible dutiliser l'interface LAN à
distance grâce au protocole de prise TCP/IP.
6
USB
Accepte un câble USB standard.
7
GPIB- En option
L’adresse GPIB par défaut est 5.
2 - Principes de fonctionnement
Disposition de l'écran
48 TGF4000 Series Quick Start Guide
Disposition de l'écran
Réf.
Brève description
Fonction
1
Indicateur de voie
Affiche la voie actuellement sélectionnée
2
Type de forme d'onde
Affiche la forme d'onde porteuse actuelle
3
État de sortie
Indique si la sortie principale est activée (On) ou désactivée (Off)
4
Témoin d'horloge externe
Affiche l'état de l'horloge externe (si demandé)
5
Indicateur d'état LAN
Affiche l'état de connexion au réseau local (Ethernet).
6
Fenêtre Paramètres
Affiche les principaux paramètres pour la forme d'onde.
7
Description du menu
Affiche le menu d'édition actuellement sélectionné.
8
Fenêtre Graphique
Affiche une représentation graphique de la forme d'onde
sélectionnée.
9
Fenêtre Modification
Affiche le paramètre actuel qui peut être modifié
10
Étiquette de touche
programmable
Affiche les fonctions actuelles pour les six touches situées en
dessous.
3 - Première utilisation
Conditions initiales
49 TGF4000 Series Quick Start Guide
3. PREMIÈRE UTILISATION
Afin de familiariser l'utilisateur avec certaines des fonctionnalités fondamentales de l'instrument,
ce guide comporte un certain nombre d'exemples de configuration.
Il est recommandé à tous les utilisateurs d'exécuter les trois premiers exemples :
Configurer un signal d'onde sinusoïdale
Configurer un signal d'horloge à onde carrée
Configuration d'une forme d'onde à impulsión
Configurer de plus amples options pour les sorties
Plusieurs autres exemples de configuration sont fournis et présument que l'utilisateur est familier
avec l'appareil :
Configurer un signal d'onde arbitraire
Configurer une forme d'onde sinusoïdale modulée AM
PRBS (TGF4162 & TGF4242)
Modulation de fréquence d'une forme d'onde sinusoïdale
Forme d'onde modulée à largeur d'impulsion (PWM)
Modulation par déplacement d
amplitude (ASK)
Balayage de fréquence d'une onde sinusoïdale
Génération d'une rafale déclenchée
Couplage des fréquences de deux voies
Compteur de fréquence
Conditions initiales
Avant de configurer l'appareil pour l'un des exemples, il doit être retourné aux conditions par
défaut.
Pour ce faire, procédez comme suit :
Appuyez sur la touche marquée Utility
Appuyez sur la touche programmable étiquetée System
Appuyez sur la touche programmable étiquetée Default (l'écran affichera Restore
Factory Default?)
Appuyez sur la touche programmable étiquetée Yes
Ceci règle la principale forme d'onde sinusoïdale (10 kHz, 1 V de crête à crête) et annule tout
déclenchement de modulations, balayages ou rafales.
REMARQUE :
L'appareil peut être réglé de façon à se souvenir de ses derniers paramètres au moment de la mise
à l'arrêt et à les restaurer lors de la mise en marche. Ce réglage peut être configuré à partir du
menu Utility avec la touche programmable PwrOn. Ce réglage sera perdu lorsque l'instrument
sera réinitialisé aux conditions par défaut, comme décrit ci-dessus.
4 - Exemples de configuration de base
Configurer un signal d'onde sinusoïdale
50 TGF4000 Series Quick Start Guide
4. EXEMPLES DE CONFIGURATION DE BASE
Configurer un signal d'onde sinusoïdale
Exigence
Émission d'un signal sinusoïdal continu d'une fréquence de 40 MHz et d'une amplitude de 6 V de
crête à crête à partir de la sortie principale 1 (MAIN OUT 1).
Conditions de démarrage
Avant de commencer, réinitialiser l'appareil aux paramètres par défaut comme décrit dans la
rubrique 3 Première utilisation
Appuyez sur la touche marquée Waves
Appuyez sur la touche programmable étiquetée Sine
4 - Exemples de configuration de base
Configurer un signal d'onde sinusoïdale
51 TGF4000 Series Quick Start Guide
Régler la fréquence
La touche programmable étiquetée Freq est mise en surbrillance : la fréquence actuelle s'affiche
dans la zone d'édition.
Notez qu'en appuyant plusieurs fois sur cette touche programmable, sa fonction bascule entre
Fréquence et Période.
Utiliser le pavé numérique pour saisir une nouvelle fréquence,
Appuyer sur les chiffres 4 0
Veuillez noter que, dès qu'un nombre est saisi, les touches programmables changent pour afficher
les unités de fréquence.
Appuyer sur la touche programmable étiquetée MHz pour confirmer une fréquence
40 MHz.
4 - Exemples de configuration de base
Configurer un signal d'onde sinusoïdale
52 TGF4000 Series Quick Start Guide
Régler l'amplitude
Appuyez sur la touche programmable étiquetée Ampl
Veuillez noter que des pressions successives sur la touche programmable Ampl changent les
étiquettes des touches Ampl et Offset qui deviennent HiLvl (niveau haut) et LoLvl (niveau bas) et
vice versa.
Utiliser le pavé numérique pour saisir une nouvelle amplitude. Appuyer sur le chiffre 6
Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher
des unités de tension.
4 - Exemples de configuration de base
Configurer un signal d'onde sinusoïdale
53 TGF4000 Series Quick Start Guide
Appuyer sur la touche programmable étiquetée Vpp pour confirmer une amplitude de
crête à crête de 6,0 V.
Activer la sortie
Appuyer sur la touche Output 1 (Sortie 1) pour activer la voie 1.
La touche de sortie 1 s'allume en orange pour indiquer l'état activé
4 - Exemples de configuration de base
Configurer un signal d'horloge à onde carrée
54 TGF4000 Series Quick Start Guide
Configurer un signal d'horloge à onde care
Exigence
Émission d'un signal d'horloge d'onde carrée continu d'une fréquence de 20 MHz, avec un cycle de
service de 50 %, un niveau élevé de 3,3 V et un niveau bas de 0,0 V provenant de la sortie
principale 1 (MAIN OUT 1).
Conditions de démarrage
Avant de commencer, réinitialiser l'appareil aux paramètres par défaut comme décrit dans la
rubrique 3 Première utilisation
Ouvrir le menu de forme d'onde - Carré
Appuyer sur la touche marquée Waves
Appuyer sur la touche programmable étiquetée Square.
4 - Exemples de configuration de base
Configurer un signal d'horloge à onde carrée
55 TGF4000 Series Quick Start Guide
Régler la fréquence
La touche programmable étiquetée Freq est mise en surbrillance : la fréquence actuelle s'affiche
dans la zone d'édition.
Notez qu'en appuyant plusieurs fois sur cette touche programmable, sa fonction bascule entre
Fréquence et Période.
Utiliser le panumérique pour saisir une nouvelle fréquence. Appuyer sur les chiffres
2 0
Veuillez noter que, dès qu'un nombre est saisi, les touches programmables changent pour afficher
les unités de fréquence.
Appuyer sur la touche programmable étiquetée MHz pour confirmer une fréquence de
20 MHz.
Notez que la fenêtre graphique change pour afficher le temps de montée sur les fronts, qui est
maintenant significatif.
4 - Exemples de configuration de base
Configurer un signal d'horloge à onde carrée
56 TGF4000 Series Quick Start Guide
Confirmer le cycle de service
Appuyer sur la touche programmable étiquetée Duty - le cycle de service actuel apparaît
dans la fenêtre d'édition.
Veuillez noter que le cycle de service est déjà réglé sur 50 %, mais qu'il peut être modifié ici le cas
échéant.
Régler les niveaux haut et bas
Appuyer sur la touche programmable étiquetée Ampl - l'étiquette de la touche devient
HiLvl et la haute de niveau haut actuelle s'affiche dans la zone d'édition
Veuillez noter que des pressions successives sur la touche programmable Ampl changent les
étiquettes des touches Ampl et Offset qui deviennent HiLvl (niveau haut) et LoLvl (niveau bas) et
vice versa.
4 - Exemples de configuration de base
Configurer un signal d'horloge à onde carrée
57 TGF4000 Series Quick Start Guide
Lorsque la touche programmable est étiquetée HiLvl - la tension de niveau haut actuelle apparaît
dans la fenêtre d'édition.
Utiliser le pavé numérique pour saisir un nouveau niveau. Appuyer sur les chiffres
3 . 3
Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher
des unités de tension.
Appuyer sur la touche programmable étiquetée V pour confirmer un niveau haut de 3.3 V.
4 - Exemples de configuration de base
Configurer un signal d'horloge à onde carrée
58 TGF4000 Series Quick Start Guide
Appuyer sur la touche programmable étiquetée LoLvl - la tension de niveau bas actuelle
apparaît dans la fenêtre d'édition.
Utiliser le pavé numérique pour saisir un nouveau niveau. Appuyer sur 0
Appuyer sur la touche programmable étiquetée V pour confirmer un niveau bas de 0,0 V.
Activer la sortie
Appuyer sur la touche Output 1 (Sortie 1) pour activer la voie 1.
La touche de sortie 1 s'allume en orange pour indiquer l'état activé
4 - Exemples de configuration de base
Configurer un signal d'horloge à onde carrée
59 TGF4000 Series Quick Start Guide
Apporter des modifications en direct à tout paramètre numérique (par ex. la
fréquence)
Les paramètres numériques peuvent être modifiés à l'aide des les touches du curseur et de la
molette au lieu d'utiliser le pavé numérique.
Appuyer sur la touche marquée Waves
Appuyer sur la touche programmable étiquetée Square.
Appuyer sur la touche programmable étiquetée Freq - la valeur de fréquence actuelle de
20 MHz s'affiche
Appuyer sur les touches du curseur pour déplacer la surbrillance d'édition jusqu'au
deuxième chiffre.
Utiliser la molette pour modifier la valeur - la fréquence change immédiatement.
REMARQUE :
Appuyez sur pour activer / désactiver la molette.
4 - Exemples de configuration de base
Configuration d'une forme d'onde à impulsión
60 TGF4000 Series Quick Start Guide
Configuration d'une forme d'onde à impulsión
Exigence
Émission d'un signal d'impulsion continu d'une période de 100 ns, avec une largeur d'impulsion de
30 ns, un temps de front de 20 ns, un niveau haut de 2,7 V et un niveau bas de -0,6 V provenant de
la sortie principale 1 (MAIN OUT 1).
Conditions de démarrage
Avant de commencer, réinitialiser l'appareil aux paramètres par défaut comme décrit dans la
rubrique 3 Première utilisation
Ouvrir le menu de forme d'onde - Impulsion
Appuyer sur la touche marquée Waves
Appuyer sur la touche programmable étiquetée Pulse.
4 - Exemples de configuration de base
Configuration d'une forme d'onde à impulsión
61 TGF4000 Series Quick Start Guide
Régler la période
Appuyer sur la touche programmable étiquetée PlsFrq de façon à ce qu'elle affiche PlsPer - la
période actuelle s'affiche dans la zone d'édition.
Notez qu'en appuyant plusieurs fois sur cette touche programmable, sa fonction bascule entre
Fréquence et Période.
Utiliser le pavé numérique pour saisir une nouvelle période. Appuyer sur les chiffres 1 0 0
Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher
des unités de temps.
4 - Exemples de configuration de base
Configuration d'une forme d'onde à impulsión
62 TGF4000 Series Quick Start Guide
Appuyer sur la touche programmable étiquetée ns pour confirmer une période de 100 ns.
Notez que la fenêtre graphique change pour afficher une représentation de l'impulsion et des
temps de front.
Régler la largeur d'impulsion
Appuye sur la touche programmable étiquetée Duty - l
étiquette de la touche devient
Width et affiche la largeur sous forme de durée
4 - Exemples de configuration de base
Configuration d'une forme d'onde à impulsión
63 TGF4000 Series Quick Start Guide
Utiliser le pavé numérique pour saisir une nouvelle largeur. Appuyer sur les chiffres 3 0.
Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher
des unités de temps.
Appuyer sur la touche programmable étiquetée ns pour confirmer une largeur de 30 ns.
4 - Exemples de configuration de base
Configuration d'une forme d'onde à impulsión
64 TGF4000 Series Quick Start Guide
Régler les temps de front d'impulsion
Appuyer sur la touche programmable étiquetée More
Appuyer sur la touche programmable étiquetée Edge
Utiliser les touches de curseur pour sélectionner le chiffre représentant des unités de
10.0 ns
Utiliser la molette pour modifier la valeur et la régler sur 20,0 ns
Veuillez noter que la valeur aurait pu être saisie à l'aide du pavé numérique selon les
préférences de chacun.
Appuyer sur la touche programmable étiquetée Done
4 - Exemples de configuration de base
Configuration d'une forme d'onde à impulsión
65 TGF4000 Series Quick Start Guide
Régler les niveaux haut et bas
Appuyer sur la touche programmable étiquetée Ampl - l'étiquette de la touche devient
HiLvl et la haute de niveau haut actuelle s'affiche dans la zone d'édition
Veuillez noter que des pressions successives sur la touche programmable Ampl changent les
étiquettes des touches Ampl et Offset qui deviennent HiLvl (niveau haut) et LoLvl (niveau bas) et
vice versa.
Utiliser le pavé numérique pour saisir un nouveau niveau. Appuyer sur les chiffres 2 . 7
Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher
des unités de tension.
4 - Exemples de configuration de base
Configuration d'une forme d'onde à impulsión
66 TGF4000 Series Quick Start Guide
Appuyer sur la touche programmable étiquetée V pour confirmer une tension haute de
2,7 V.
Appuyer sur la touche programmable étiquetée LoLvl - la tension de niveau bas actuelle
apparaît dans la fenêtre d'édition.
4 - Exemples de configuration de base
Configuration d'une forme d'onde à impulsión
67 TGF4000 Series Quick Start Guide
Utiliser le pavé numérique pour saisir un nouveau niveau. Appuyer sur - . 6
Appuyer sur la touche programmable étiquetée V pour confirmer un niveau bas de -600 mV.
Activer la sortie
Appuyer sur la touche Output1 (Sortie 1) pour activer la voie 1.
La touche de sortie 1 s'allume en orange pour indiquer l'état activé
4 - Exemples de configuration de base
Configurer de plus amples options pour les sorties
68 TGF4000 Series Quick Start Guide
Configurer de plus amples options pour les sorties
Exigence
Les exemples de configuration précédents ont montré comment utiliser le menu sortie pour définir
le niveau de sortie (amplitude plus décalage ou niveau haut au plus niveau bas) et pour activer ou
désactiver la sortie. Cet exemple présente le réglage de la phase de sortie, de la polarité de sortie,
de l'impédance de charge et de la plage automatique de tension.
Conditions de démarrage
Avant de commencer, réinitialiser l'appareil aux paramètres par défaut comme décrit dans la
rubrique 3 Première utilisation
Ouvrir le Menu Output (Sortie)
Appuyer sur la touche programmable marquée Output
4 - Exemples de configuration de base
Configurer de plus amples options pour les sorties
69 TGF4000 Series Quick Start Guide
Modifier la phase de sortie
La touche programmable étiquetée Phase est sélectionnée par défaut.
Saisir une phase de -45 degrés.
L'angle de phase réglé est le point de la période de la forme d'onde qui coïncide avec le front de
synchronisation ou de déclenchement. Autrement dit, il s'agit du point de la période auquel
commence la forme d'onde. C'est pourquoi un réglage de phase négatif avance, et un réglage de
phase positif retarde la forme d'onde par rapport à la synchronisation ou au déclenchement ; la
forme d'onde présentée dans la fenêtre graphique change pour illustrer cet état de fait.
La touche programmable Align est utilisée pour réaligner la phase lorsque des modifications sont
apportées à la fréquence.
Modifier la polarité de sortie
Appuyer sur la touche programmable étiquetée Type pour inverser la polarité de sortie.
Veuillez noter que des pressions successives de la touche Type permettent d'alterner entre une
polarité normale et inversée.
4 - Exemples de configuration de base
Configurer de plus amples options pour les sorties
70 TGF4000 Series Quick Start Guide
Modifier l'impédance de charge
Appuyez sur la touche programmable étiquetée Load
Appuyer à nouveau sur la touche programmable étiquetée Load pour modifier l'impédance
de charge et la régler sur High-z. (Impédance élevée).
Veuillez noter que, par défaut, l'impédance de charge est de 50 Ohms, mais que cette valeur peut
être modifiée au profit de n'importe quelle impédance d'une valeur comprise entre 50 et
10 000 Ohms. Les niveaux sont calculés en fonction de cette impédance.
Des pressions successives de la touche Load permettent d'alterner entre une valeur numérique et
High-z. Veuillez noter que la mesure de l'amplitude augmente pour atteindre 2 V de crête à crête.
Appuyer sur la touche programmable étiquetée Load pour ramener l'impédance de charge
à 50 Ohms.
4 - Exemples de configuration de base
Configurer de plus amples options pour les sorties
71 TGF4000 Series Quick Start Guide
Modifier la plage
Appuyez sur la touche programmable étiquetée range
Appuyer à nouveau sur la touche programmable étiquetée range pour faire passer la plage
du mode Auto (automatique) au mode Hold (maintien).
Le mode Auto règle la gamme automatiquement par pas de 6 dB d
atténuateur (c.-à-d. gamme
maximum de 10 Vpp, 5 Vpp, 2,5 Vpp, etc., à 50 avec une gamme du vernier de réglage de
l
amplitude limitée à 6 dB afin de maintenir la qualité de la forme donde.
La sélection du mode Hold désactive le réglage automatique de la gamme ; le réglage actuel de
l
atténuateur est fixé et la gamme du vernier de réglage de lamplitude nest plus limitée.
Lorsque la gamme est réglée sur Auto, l'amplitude et les atténuateurs sont automatiquement
activés pour obtenir une performance optimale. Lorsque la plage est réglée sur Hold (Maintien), un
réglage fixe de l
atténuateur est utilisé pour tous les réglages de lamplitude.
5 - Explorer les fonctionnalités du générateur
Configurer un signal d'onde arbitraire
72 TGF4000 Series Quick Start Guide
5. EXPLORER LES FONCTIONNALITES DU
GENERATEUR
Dans les exemples suivants, seuls les réglages des paramètres sont présentés. Ils sont accompagnés
des noms des touches correspondantes. Les formes d'onde de sortie ainsi obtenues sont affichées,
ainsi que la forme d'onde de synchronisation ou de déclenchement le cas échéant. Les paramètres
d'amplitude et de décalage de la sortie sont fournis à titre illustratif uniquement et ne doivent pas
nécessairement être adoptés.
Configurer un signal d'onde arbitraire
Commencer avec l'instrument réinitialisé aux paramètres par défaut.
MENU
Nom de la touche programmable
D'ondes
Waves
MENU
NOM DE LA TOUCHE
Forme donde arbitraire
Arb (Arbitraire)
Paramètre
Nom de la touche programmable
Réglage
Sélection de londe
Waves (Ondes)
Sinc
Fréquence
-
10 kHz
Amplitude
Ampl
2 V
Décalage
Offset (Décalage)
5 mV
Paramètre
NOM DE LA TOUCHE
Réglage
État de sortie
Output1 (Sortie 1)
On
5 - Explorer les fonctionnalités du générateur
Configurer une forme d'onde sinusoïdale modulée AM
73 TGF4000 Series Quick Start Guide
Configurer une forme d'onde sinusoïdale modulée AM
Commencer avec l'instrument réinitialisé aux paramètres par défaut. .
Paramètre
Nom de la touche programmable
Réglage
Fréquence
-
10 MHz
MENU
NOM DE LA TOUCHE
Modulation
Mod
Paramètre
Nom de la touche programmable
Réglage
Fréquence
-
100 kHz
Profondeur
Depth (Profondeur)
100 %
Source
Source
Internal
Forme
Shape (Forme)
Sine
État de modulation
On/Off
On
Paramètre
NOM DE LA TOUCHE
Réglage
État de sortie
Output1 (Sortie 1)
On
Changer la modulation et la régler sur AM-SC
Paramètre
Nom de la touche programmable
Réglage
Type
Type > AM
AM-SC
5 - Explorer les fonctionnalités du générateur
PRBS
74 TGF4000 Series Quick Start Guide
PRBS
Commencer avec l'instrument réinitialisé aux paramètres par défaut.
MENU
NOM DE LA TOUCHE
D'ondes
Waves
MENU
Nom de la touche programmable
Réglage
Bruit/PRBS
Noise (Bruit)
Source
Source
PBRS
Paramètre
Nom de la touche programmable
Réglage
Débit binaire
BitRate
1 Mbps
Amplitude
Ampl
3,3 V
Décalage
Décalage
1,65 V
Type de PRBS
Type
PN7
Paramètre
NOM DE LA TOUCHE
Réglage
État de sortie
Output1 (Sortie 1)
On
5 - Explorer les fonctionnalités du générateur
Modulation de fréquence d'une forme d'onde sinusoïdale
75 TGF4000 Series Quick Start Guide
Modulation de fréquence d'une forme d'onde sinusoïdale
Commencer avec l'instrument réinitialisé aux paramètres par défaut.
MENU
NOM DE LA TOUCHE
Modulation
Mod
Paramètre
Nom de la touche programmable
Réglage
État de modulation
On/Off
On
Type de modulation
Type
FM
Fréquence de modulation
-
1 kHz
Déviation
Deviatn
9 kHz
MENU
NOM DE LA TOUCHE
Sinusoïde
Params
Paramètre
Nom de la touche programmable
Réglage
Amplitude
Ampl
1,0 V
Décalage
Offset (Décalage)
0,0 V
Paramètre
NOM DE LA TOUCHE
Réglage
État de sortie
Output1 (Sortie 1)
On
5 - Explorer les fonctionnalités du générateur
Forme d'onde modulée à largeur d'impulsion (PWM)
76 TGF4000 Series Quick Start Guide
Forme d'onde modulée à largeur d'impulsion (PWM)
Commencer avec l'instrument réinitialisé aux paramètres par défaut.
MENU
NOM DE LA TOUCHE
D'ondes
Waves
Paramètre
Nom de la touche programmable
Réglage
Impulsion
Pulse
MENU
NOM DE LA TOUCHE
Modulation
Mod
Paramètre
Nom de la touche programmable
Réglage
État de modulation
On/Off
On
Type de modulation
Type
PWM
Fréquence de
modulation
-
1 kHz
Déviation
Dev %
40 %
MENU
NOM DE LA TOUCHE
Impulsion
Params
Paramètre
Nom de la touche programmable
Réglage
Amplitude
Ampl
1,0 V
Décalage
Décalage
0,0 V
Paramètre
NOM DE LA TOUCHE
Réglage
État de sortie
Output1 (Sortie 1)
On
5 - Explorer les fonctionnalités du générateur
Modulation par déplacement damplitude (ASK)
77 TGF4000 Series Quick Start Guide
Modulation par déplacement damplitude (ASK)
Commencer avec l'instrument réinitialisé aux paramètres par défaut.
MENU
NOM DE LA TOUCHE
Modulation
Mod
Paramètre
Nom de la touche programmable
Réglage
État de modulation
On/Off
On
Type de modulation
Type
ASK
Source de modulation
Source
Internal
Amplitude de saut
HpAmpl
100 mV
Taux de commutation
Rate (Taux)
1 kHz
Polarité de saut
HopPol
Positive
MENU
NOM DE LA TOUCHE
Sinusoïde
Params
Paramètre
Nom de la touche programmable
Réglage
Amplitude
Ampl
1,0 V
Décalage
Décalage
0,0 V
Paramètre
NOM DE LA TOUCHE
Réglage
État de sortie
Output1 (Sortie 1 )
On
Notez que le deuxième tracé est la sortie de la prise Output 2 (Sortie 2) lorsque la voie 2 est
synchronisée, qui suit le signal de la forme d'onde modulée.
5 - Explorer les fonctionnalités du générateur
Balayage de fréquence d'une onde sinusoïdale
78 TGF4000 Series Quick Start Guide
Balayage de fréquence d'une onde sinusoïdale
Commencer avec l'instrument réinitialisé aux paramètres par défaut.
MENU
NOM DE LA TOUCHE
Balayage
Sweep
Paramètre
Nom de la touche programmable
Réglage
État de balayage
On/Off
On
Fréquence darrêt
Freq > Stop
100 kHz
MENU
NOM DE LA TOUCHE
Sinusoïde
Params
Paramètre
Nom de la touche programmable
Réglage
Amplitude
Ampl
1,0 V
Décalage
Décalage
0,0 V
Paramètre
NOM DE LA TOUCHE
Réglage
État de sortie
Output1 (Sortie 1)
On
5 - Explorer les fonctionnalités du générateur
Génération d'une rafale déclenchée
79 TGF4000 Series Quick Start Guide
Génération d'une rafale déclenchée
Commencer avec l'instrument réinitialisé aux paramètres par défaut.
Paramètre
Nom de la touche programmable
Réglage
Fréquence
-
6 MHz
MENU
NOM DE LA TOUCHE
Rafales
Burst
Paramètre
Nom de la touche programmable
Réglage
État des rafales
On/Off
On
Compte de rafales
Count
3
Source de déclenchement
SetTrg > Source > Int
Internal
Période de déclenchement
SetTrg > Period
5 ms
MENU
NOM DE LA TOUCHE
Sinusoïde
Params
Paramètre
Nom de la touche programmable
Réglage
Amplitude
Ampl
1,0 V
Décalage
Décalage
0,0 V
Paramètre
NOM DE LA TOUCHE
Réglage
État de sortie
Output1 (Sortie 1)
On
Notez que le deuxième tracé est la sortie de la prise Main Output 2 (Sortie principale 2) lorsque la
voie 2 est synchronisée, qui suit le signal d
entrée de déclenchement.
5 - Explorer les fonctionnalités du générateur
Couplage des fréquences de deux voies
80 TGF4000 Series Quick Start Guide
Couplage des fréquences de deux voies
Commencer avec l'instrument réinitialisé aux paramètres par défaut.
MENU
NOM DE LA TOUCHE
Utilitaires
Utility
Paramètre
Nom de la touche programmable
Réglage
Fréquences
Dual Ch >Freq
Coupled
Paramètre
NOM DE LA TOUCHE
Réglage
État de sortie
Output1
On
État de sortie
Output2
On
MENU
NOM DE LA TOUCHE
Sinusoïde
Params
Paramètre
Nom de la touche programmable
Réglage
Changement de phase
Output > Phase > Done
90 degrés
Fréquence
Freq
1MHz
Veuillez noter que lorsque la voie 1 est réglée sur 1MHz, la voie 2 est également réglée sur 1MHz.Il
est possible d'observer le changement de phase de 90 degrés entre les voies.
5 - Explorer les fonctionnalités du générateur
Compteur de fréquence
81 TGF4000 Series Quick Start Guide
Compteur de fréquence
Commencer avec l'instrument réinitialisé aux paramètres par défaut.
MENU
NOM DE LA TOUCHE
Utilitaires
Utility
Paramètre
Nom de la touche programmable
Réglage
Compteur
Instr > FrCntr > On/Off
Counter Enabled
Source
Source
TRIG IN - DC Coupled
Type
Freq
Frequency
Mesure
Count
La fenêtre Modification affiche la mesure actuelle. Sans la présence d
un signal dentrée à lentrée
sélectionnée, le compteur affiche « No signal » jusqu'à ce qu'un signal d'entrée adéquat soit
appliqué.
Lorsqu
un signal dentrée adéquat est appliqué, le compteur mesure et affiche constamment la
mesure dans la case Edit.
6 - Maintenance
Nettoyage
82 TGF4000 Series Quick Start Guide
6. MAINTENANCE
Le fabricant ou ses agences à létranger proposent un service de réparation pour tout appareil
défaillant. Si les propriétaires de ce matériel souhaitent en effectuer l
entretien par leurs propres
moyens, il est fortement recommandé que ce travail soit effectué exclusivement par un personnel
qualifié, à l
aide du guide dentretien que lon peut acheter directement chez le fabricant ou ses agents à
l
étranger.
Nettoyage
Si lappareil a besoin dêtre nettoyé, utiliser uniquement un chiffon qui aura été préalablement
humidifié avec un peu d
eau ou un détergent doux.
AVERTISSEMENT ! AFIN DÉVITER TOUT RISQUE DÉLECTROCUTION OU UNE DÉTÉRIORATION DE
LAPPAREIL, NE JAMAIS LAISSER DEAU SINFILTRER À LINTÉRIEUR DU BOÎTIER. POUR ÉVITER TOUTE
DÉTÉRIORATION, NE JAMAIS UTILISER DE SOLVANTS POUR NETTOYER LE BOÎTIER.
0 - Inhaltsverzeichnis
83 TGF4000 Series Quick Start Guide
INHALTSVERZEICHNIS
 Sicherheit ........................................................................................................................................................ 84
Symbole ............................................................................................................................................................. 85
 Funktionsprinzipien ...................................................................................................................................... 86
Layout des vorderen Bedienfelds ...................................................................................................................... 86
Buchsen auf der Rückseite ................................................................................................................................ 87
Bildschirmaufbau .............................................................................................................................................. 88
 Erste Schritte ................................................................................................................................................ 89
Anfangsbedingungen......................................................................................................................................... 89
 Grundlegende Einrichtungsbeispiele ............................................................................................................ 90
Einrichten eines Sinussignals ............................................................................................................................. 90
Einrichten eines Rechteck-Taktsignals ............................................................................................................... 94
Einrichten einer Impulswellenform ................................................................................................................. 100
Einrichten weiterer Ausgangsoptionen ........................................................................................................... 108
 Erkunden der Generator-Funktionen .......................................................................................................... 112
Einrichten einer Arbiträr-Wellenform .............................................................................................................. 112
Einrichten einer AM-modulierten Sinuskurve ................................................................................................. 112
PRBS ................................................................................................................................................................ 114
Frequenzmodulation einer Sinuskurve ............................................................................................................ 114
Pulsweitenmodulierte Wellenform (PWM) ..................................................................................................... 116
Amplitudenumtastung (ASK) ........................................................................................................................... 117
Frequenz-Sweep einer Sinuswelle ................................................................................................................... 118
Generieren eines getriggerten Burst-Signals ................................................................................................... 119
Frequenzkopplung beider Kanäle .................................................................................................................... 120
Frequenzzähler ................................................................................................................................................ 121
 Wartung ..................................................................................................................................................... 122
Reinigung ........................................................................................................................................................ 122
Die neuesten Versionen dieses Handbuchs, Gerätetreiber und Softwaretools können von folgender
Website heruntergeladen werden: www.aimtti.com
1 - Sicherheit
84 TGF4000 Series Quick Start Guide
1. SICHERHEIT
Dieses Gerät wurde nach der Sicherheitsklasse (Schutzart) I der IEC-Klassifikation und gemäß den
europäischen Vorschriften EN61010-1 (Sicherheitsvorschriften für Elektrische Meß-, Steuer, Regel-
und Laboranlagen) entwickelt. Es handelt sich um ein Gerät der Installationskategorie II, das für den
Betrieb von einer normalen einphasigen Versorgung vorgesehen ist.
Das Gerät wurde gemäß den Vorschriften EN61010-1 geprüft und wurde in sicherem Zustand
geliefert. Die vorliegende Anleitung enthält vom Benutzer zu beachtende Informationen und
Warnungen, die den sicheren Betrieb und den sicheren Zustand des Gerätes gewährleisten.
Dieses Gerät ist für den Betrieb in Innenräumen der Umgebungsklass 2 , für einen
Temperaturbereich von 5° C bis 40° C und 20 - 80 % relative Feuchtigkeit (nicht kondensierend)
vorgesehen. Gelegentlich kann es Temperaturen zwischen +5° und 10°C ausgesetzt sein, ohne d
seine Sicherheit dadurch beeinträchtigt wird. Betreiben Sie das Gerät jedoch auf keinen Fall,
solange Kondensation vorhanden ist.
Ein Einsatz dieses Geräts in einer Weise, die für diese Anlage nicht vorgesehen ist, kann die
vorgesehene Sicherheit beeinträchtigen. Auf keinen Fall das Gerät außerhalb der angegebenen
Nennversorgungsspannungen oder Umgebungsbedingungen betreiben.
WARNUNG! - DIESES GERÄT MUSS GEERDET WERDEN!
Jede Unterbrechung des Netzschutzleiters innerhalb oder außerhalb des Geräts macht das Gerät
gefährlich. Eine absichtliche Unterbrechung ist verboten. Die Schutzwirkung darf durch
Verwendung eines Verlängerungskabels ohne Schutzleiter nicht aufgehoben werden.
Ist das Gerät an die elektrische Versorgung angeschlossen, so können die Klemmen unter Spannung
stehen, was bedeutet, daß beim Entfernen von Verkleidungs- oder sonstigen Teilen (mit Ausnahme
der Teile, zu denen Zugang mit der Hand möglich ist) höchstwahrscheinlich spannungsführende
Teile bloßgelegt weden. Vor jeglichem Öffnen des Geräts zu Nachstell-, Auswechsel-, Wartungs-
oder Reparaturzwecken, Gerät stets von sämtlichen Spannungsquellen abklemmen.
Jegliche Nachstellung, Wartung und Reparatur am geöffneten, unter Spannung stehenden Gerät,
ist nach Möglichkeit zu vermeiden. Falls unvermeidlich, sollten solche Arbeiten nur von
qualifiziertem Personal ausgeführt werden, das sich der Gefahren bewußt ist.
Ist das Gerät eindeutig fehlberbehaftet, bzw. wurde es mechanisch beschädigt, übermäßiger
Feuchtigkeit oder chemischer Korrosion ausgesetzt, so können die Schutzeinrichtungen
beeinträchtigt sein, weshalb das Gerät aus dem Verkehr zurückgezogen und zur Überprüfung und
Reparatur eingesandt werden sollte.
Sicherstellen, daß nur Sicherungen der vorgeschriebenen Stromstärke und des vorgesehenen Typs
als Ersatz verwendet werden. Provisorische “Sicherungenund der Kurzschluß von
Sicherungshaltern ist verboten.
Zur Batteriepufferung des Speichers wird bei diesem Gerät eine Lithium-Knopfzelle verwendet. Ihre
typische Lebensdauer beträgt 5 Jahre. Wenn sie ausgewechselt werden muß, darf sie nur durch
eine Zelle desselben Typs ersetzt werden, d.h. eine 3 V Li/MnO
2
20 mm Knopfzelle, Typ 2032.
Verbrauchte Zellen müssen entsprechend der am Ort geltenden Vorschriften entsorgt werden. Auf
keinen Fall darf versucht werden, die Zelle aufzuladen, zu öffnen, zu verbrennen, oder sie
Temperaturen von über 60° C auszusetzen.
1 - Sicherheit
Symbole
85 TGF4000 Series Quick Start Guide
Symbole
Beim Reinigen darauf achten, daß das Gerät nicht naß wird. Am Gerät werden folgende Symbole
verwendet:
Vorsicht!
Bitte beachten Sie die beigefügten Unterlagen.
Falsche Bedienung kann Schaden am Get verursachen!
Masse
Wechselstrom
Netz OFF (aus)
l
Netz ON (ein)
2 - Funktionsprinzipien
Layout des vorderen Bedienfelds
86 TGF4000 Series Quick Start Guide
2. FUNKTIONSPRINZIPIEN
Layout des vorderen Bedienfelds
Ref Kurzbeschreibung Funktion
1 Stromschalter
Schaltet das Gerät ein oder aus. Sicherheitshinweis: Ziehen
Sie den Netzstecker an der Geräterückseite oder schalten
Sie die Netzsteckdose aus, um das Gerät ganz vom Netz zu
trennen. Achten Sie darauf, dass die Abschaltmöglichkeit
gut zugänglich ist.
2 Softkeys Führt die Funktion auf dem LCD-Softkey aus.
3 Wellenform-Menüs
Wählt den Hauptwellentyp (Trägersignal). (Sinus, Rechteck,
Rampe, Puls, Rausch/PBRS oder Arb.)
4
Menüs für die Wellenform-
Modifizierung
Öffnet die Menüs zum Einstellen der Parameter für
Modulation, Sweep und Burst
5 Andere Menüs
Wählt Menüs für interne und externe Speicherung von
Dateien, Dienstprogramme und Triggerbedingungen.
6 Hauptbuchsen
Hauptausgangsbuchsen. Kanal 2 kann auch so konfiguriert
werden, dass das Kanal 1 Sync-Signal an der MAIN OUT 2-
Buchse ausgegeben wird.
7 Ausgangstasten Den gewählten MAIN OUT ein- oder ausschalten.
8 Cursor-Tasten und Drehgeber
Zur Änderung numerischer Parameterwerte Ziffer für Ziffer.
Zur Auswahl bestimmter Elemente innerhalb einiger
Menüs.
9
Numerische Tastatur
Zur direkten Eingabe numerischer Parameterwerte.
10
USB-Stick
USB-Host-Anschluss für USB-Flash-Laufwerk.
2 - Funktionsprinzipien
Buchsen auf der Rückseite
87 TGF4000 Series Quick Start Guide
Buchsen auf der Rückseite
Ref
Kurzbeschreibung
Funktion
1
Modulationseingang
Eingang zur externen Modulation der Hauptwellenformen.
2
Referenz Eing. / AC-
gekoppelter Frequenzzähler
Eingang für externen 10MHz Referenztakt und AC gekoppelte
externe Frequenzmessung.
3
Referenz Ausg.
Ausgang für internes 10 MHz Referenztaktsignal.
4
Trigger-Eingang / DC-
gekoppelter Frequenzzähler
Eingang für externe Triggerung von Hauptwellenformen und DC-
gekoppelte externe Frequenzmessung
5 LAN-Schnittstelle
Die LAN-Schnittstelle entspricht LXI ( Lan eXtensions for
Instrumentation) Core 2011. Eine Fernsteuerung über die LAN-
Schnittstelle ist mittels TCP/IP Socket-Protokoll möglich.
6 USB-Port Akzeptiert ein Standard-USB-Kabel.
7
GPIB (IEEE488) -wahlweise
Die Standard-GBIP-Adresse ist 5.
2 - Funktionsprinzipien
Bildschirmaufbau
88 TGF4000 Series Quick Start Guide
Bildschirmaufbau
Ref
Kurzbeschreibung
Funktion
1
Kanalanzeige
Zeigt momentan ausgewählten Kanal
2
Hauptwellenformtyp
Zeigt das aktuelle Trägersignal
3
Ausgangsstatus
Einschalten, Ausgang Ein/Aus
4 Externer Takt-Anzeige
Zeigt den Status des externen Taktgebers an (falls
vorhanden)
5
LAN Statusanzeige
Zeigt den Status des LAN-Anschluss an (Ethernet).
6
Parameterfeld
Zeigt die wichtigsten Parameter der Wellenform.
7
Menübeschreibung
Zeigt das aktuelle gewählte Bearbeitungsmenü.
8 Grafikbereich
Zeigt eine graphische Darstellung der ausgewählten
Wellenform.
9
Bearbeitungsfeld
Zeigt den aktuellen Parameter, der bearbeitet werden kann
10
Softkey-Label
Zeigt die aktuellen Funktionen für die sechs Tasten unten.
3 - Erste Schritte
Anfangsbedingungen
89 TGF4000 Series Quick Start Guide
3. ERSTE SCHRITTE
Um den Benutzer mit einigen der grundlegenden Funktionen des Instruments vertraut zu machen,
werden hier eine Reihe von Setup Beispielen beschrieben.
Es wird empfohlen, dass alle Benutzer zumindest die ersten drei Beispiele ausführen sollten:
Einrichten eines Sinussignals
Einrichten eines Rechteck-Taktsignals
Einrichten einer Impulswellenform
Einrichten weiterer Ausgangsoptionen
Eine Reihe weiterer Set-up Beispiele werden zur Verfügung gestellt, die eine gewisse Vertrautheit
mit dem Gerät voraussetzen:
Einrichten einer Arbiträr-Wellenform
Einrichten einer AM-modulierten Sinuskurve
PRBS (TGF4162 & TGF4242)
Frequenzmodulation einer Sinuskurve
Pulsweitenmodulierte Wellenform (PWM)
Amplitudenumtastung (ASK)
Frequenz-Sweep einer Sinuswelle
Generieren eines getriggerten Burst-Signals
Frequenzkopplung beider Kanäle
Frequenzzähler
Anfangsbedingungen
Vor dem Einrichten des Geräts für die Beispiele sollte es auf Standardbedingungen zurückgesetzt
werden.
Dies geschieht folgendermaßen:
Drücken Sie die Taste mit der Bezeichnung Utility
Drücken Sie den Softkey mit der Bezeichnung System
Drücken Sie den Softkey mit der Bezeichnung Default (Display zeigt Restore Factory
Default?)
Drücken Sie den Softkey Yes
Dies stellt die Hauptwellenform auf Sinus (10 kHz, 1 V pk-pk) ein und stoppt alle Modulationen,
Sweeps, Burst-Triggers oder Gating.
ANMERKUNG:
Das Gerät kann so eingestellt werden, dass es seine letzten Einstellungen beim Ausschalten
speichert und diese beim Einschalten wiederherstellt. Dies wird im Utility-Menü und über den
PwrOn
Softkey eingestellt. Diese Einstellung geht verloren, wenn das Gerät auf die
Ausgangsbedingungen zurückgesetzt wird (siehe oben).
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Sinussignals
90 TGF4000 Series Quick Start Guide
4. GRUNDLEGENDE EINRICHTUNGSBEISPIELE
Einrichten eines Sinussignals
Voraussetzung
Ausgabe eines kontinuierlichen Sinuswellensignals mit 40 MHz Frequenz und einer Amplitude von
6 V pk-pk an MAIN OUT 1.
Startbedingungen
Bevor Sie beginnen, sollten Sie das Gerät auf Standardwerte zurücksetzen, wie in Abschnitt
Anfangsbedingungen
Öffnen Sie das Wellenform-Menü - Sine (Sinus)
Drücken Sie die Taste mit der Bezeichnung Waves
Drücken Sie den Softkey Sine
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Sinussignals
91 TGF4000 Series Quick Start Guide
Frequenzeinstellung
Der Softkey Freq wird markiert - die aktuelle Frequenz erscheint im Eingabefeld.
Beachten Sie, dass durch wiederholtes Drücken dieses Softkeys die Funktion zwischen Frequenz
und Periode wechselt.
Geben Sie eine neue Frequenz mit der Nummerntastatur ein. Drücken Sie die Ziffern 4 0
Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Frequenzeinheiten an.
Drücken Sie den Softkey MHz, um eine Frequenz von 40 MHz zu bestätigen.
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Sinussignals
92 TGF4000 Series Quick Start Guide
Amplitude festlegen
Drücken Sie den Softkey Ampl
Beachten Sie, dass durch wiederholtes Drücken des Softkeys Ampl die Beschriftung der Tasten
Ampl und Offset auf HiLvl (High-Level) und LoLvl (Low-Level) wechselt und umgekehrt.
Geben Sie eine neue Amplitude mit der Nummerntastatur ein. Drücken Sie die Ziffer 6
Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Spannungseinheiten an.
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Sinussignals
93 TGF4000 Series Quick Start Guide
Drücken Sie den Softkey Vpp, um einen pk-pk Pegel von 6,0 Volt zu bestätigen.
Schalten Sie den Ausgang ein
Drücken Sie die Taste Output 1 um den Kanal 1-Ausgang einzuschalten.
Die Ausgangstaste 1 leuchtet orange, um den Einschaltzustand anzuzeigen
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Rechteck-Taktsignals
94 TGF4000 Series Quick Start Guide
Einrichten eines Rechteck-Taktsignals
Voraussetzung
Ausgabe eines kontinuierlichen Rechtecktaktsignals mit 20 MHz Frequenz, Tastverhältnis von 50 %
und einem max. Pegel von 3,3 V und min. Pegel von 0,0 Volt an MAIN OUT 1.
Startbedingungen
Bevor Sie beginnen, sollten Sie das Gerät auf Standardwerte zurücksetzen, wie in Abschnitt
Anfangsbedingungen
Öffnen Sie das Wellenform-Menü - Square (Rechteck)
Drücken Sie die Taste mit der Bezeichnung Waves
Drücken Sie den Softkey Square.
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Rechteck-Taktsignals
95 TGF4000 Series Quick Start Guide
Frequenzeinstellung
Der Softkey Freq wird hervorgehoben - die aktuelle Frequenz erscheint im Eingabefeld.
Beachten Sie, dass durch wiederholtes Drücken dieses Softkeys die Funktion zwischen Frequenz
und Periode wechselt.
Geben Sie eine neue Frequenz mit der Nummerntastatur ein. Drücken Sie die Ziffern 2 0
Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Frequenzeinheiten an.
Drücken Sie den Softkey MHz um eine Frequenz von 20 MHz zu bestätigen.
Beachten Sie, dass die Änderungen im Grafikbereich die Anstiegszeit an den Flanken zeigen, die
jetzt von Bedeutung ist.
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Rechteck-Taktsignals
96 TGF4000 Series Quick Start Guide
Bestätigen Sie das Tastverhältnis
Drücken Sie den Softkey
Duty - das aktuelle Tastverhältnis erscheint im Eingabefeld.
Beachten Sie, dass das Tastverhältnis bereits auf 50 % festgelegt ist, jedoch hier ggf. geändert
werden kann.
Einstellen der High und Low Level
Drücken Sie den Softkey Ampl - die Tastenbeschriftung wechselt auf HiLvl und die aktuelle
Spannung mit hohem Pegel erscheint im Bearbeitungsfeld
Beachten Sie, dass durch wiederholtes Drücken des Softkeys Ampl die Beschriftung der Tasten
Ampl und Offset auf HiLvl (High-Level) und LoLvl (Low-Level) wechselt und umgekehrt.
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Rechteck-Taktsignals
97 TGF4000 Series Quick Start Guide
Wenn auf dem Softkey HiLvl angegeben ist, erscheint der aktuelle High-Level Spannungspegel im
Eingabefeld.
Geben Sie einen neuen Pegelwert mit der Nummerntastatur ein. Drücken Sie die Ziffern
3 . 3
Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Spannungseinheiten an.
Drücken Sie den Softkey V, um einen High-Pegel von 3,3 Volt zu bestätigen.
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Rechteck-Taktsignals
98 TGF4000 Series Quick Start Guide
Drücken Sie den Softkey LoLvl - der aktuelle Low-Level Spannungspegel erscheint im
Eingabefeld.
Geben Sie einen neuen Pegelwert mit der Nummerntastatur ein. Drücken Sie 0
Drücken Sie den Softkey V um einen Low-Pegel von 0,0 Volt zu bestätigen.
Schalten Sie den Ausgang ein
Drücken Sie die Taste Output1, um den Kanal 1-Ausgang einzuschalten.
Die Ausgangstaste 1 leuchtet orange, um den Einschaltzustand anzuzeigen
4 - Grundlegende Einrichtungsbeispiele
Einrichten eines Rechteck-Taktsignals
99 TGF4000 Series Quick Start Guide
Laufende Änderungen an numerischen Parametern vornehmen (z. B.
Frequenz)
Numerische Parameter können mit den Cursortasten und dem Drehgeber alternativ zum
numerischen Tastenfeld geändert werden.
Drücken Sie die Taste mit der Bezeichnung Waves.
Drücken Sie den Softkey Square.
Drücken Sie den Softkey Freq - der aktuelle Frequenzwert von 20,0 MHz wird angezeigt
Drücken Sie die Cursortasten, um die Bearbeitungsmarkierung auf die zweite Stelle zu
setzen.
Verändern Sie den Wert mit dem Drehgeber die Frequenz wird sofort geändert.
ANMERKUNG:
Drücken Sie, um das Drehgeber zu aktivieren / deaktivieren.
4 - Grundlegende Einrichtungsbeispiele
Einrichten einer Impulswellenform
100 TGF4000 Series Quick Start Guide
Einrichten einer Impulswellenform
Voraussetzung
Ausgabe eines kontinuierlichen Impulssignals mit 100 ns Periode, 30 ns Pulsbreite, 20 ns
Flankenzeiten, einem High Pegel von 2,7 V und einem Low Pegel von -0,6 Volt an MAIN OUT 1.
Startbedingungen
Bevor Sie beginnen, sollten Sie das Gerät auf Standardwerte zurücksetzen, wie in Abschnitt
Anfangsbedingungen
Öffnen Sie das Wellenform-Menü - Pulse (Impuls)
Drücken Sie die Taste mit der Bezeichnung Waves.
Drücken Sie den Softkey Pulse.
4 - Grundlegende Einrichtungsbeispiele
Einrichten einer Impulswellenform
101 TGF4000 Series Quick Start Guide
Periode festlegen
Drücken Sie den Softkey PlsFrq so dass dieser auf PlsPer wechselt - die aktuelle Periode
erscheint im Eingabefeld.
Durch wiederholtes Drücken dieses Softkeys die Funktion zwischen Frequenz und Periode wechselt.
Geben Sie einen neuen Periodenwert mit der Nummerntastatur ein. Drücken Sie die Ziffern
1 0 0
Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Zeiteinheiten an.
4 - Grundlegende Einrichtungsbeispiele
Einrichten einer Impulswellenform
102 TGF4000 Series Quick Start Guide
Drücken Sie den Softkey ns, um eine Periode von 100 ns zu bestätigen.
Im Grafikbereich erscheint nun eine Darstellung der Impuls- und Flankenzeiten.
Impulsbreite festlegen
Drücken Sie den Softkey Duty - die Tastenbeschriftung wechselt auf Width und zeigt die
Breite als Zeit an
4 - Grundlegende Einrichtungsbeispiele
Einrichten einer Impulswellenform
103 TGF4000 Series Quick Start Guide
Geben Sie einen neuen Breitenwert mit der Nummerntastatur ein. Drücken Sie die Ziffern
3 0
Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Zeiteinheiten an.
Drücken Sie den Softkey ns, um eine Breite von 30 ns zu bestätigen.
4 - Grundlegende Einrichtungsbeispiele
Einrichten einer Impulswellenform
104 TGF4000 Series Quick Start Guide
Impulsflankenzeiten festlegen
Drücken Sie den Softkey More
Drücken Sie den Softkey Edge
Wählen Sie mit den Cursortasten die Ziffer, die Einheiten von 10 ns repräsentiert
Ändern Sie den Wert mit dem Drehgeber auf 20,0 ns
Sie hätten den Wert aber auch über die numerische Tastatur eingeben können.
Drücken Sie den Softkey Done.
4 - Grundlegende Einrichtungsbeispiele
Einrichten einer Impulswellenform
105 TGF4000 Series Quick Start Guide
Einstellen der High und Low Level
Drücken Sie den Softkey Ampl - die Tastenbeschriftung wechselt auf HiLvl und die aktuelle
Spannung mit hohem Pegel erscheint im Bearbeitungsfeld
Beachten Sie, dass durch wiederholtes Drücken des Softkeys Ampl die Beschriftung der Tasten
Ampl und Offset auf HiLvl (High-Level) und LoLvl (Low-Level) wechselt und umgekehrt.
Geben Sie einen neuen Pegelwert mit der Nummerntastatur ein. Drücken Sie die Ziffern
2 . 7.
Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Spannungseinheiten an.
4 - Grundlegende Einrichtungsbeispiele
Einrichten einer Impulswellenform
106 TGF4000 Series Quick Start Guide
Drücken Sie den Softkey V, um einen High-Pegel von 2,7 Volt zu bestätigen.
Drücken Sie den Softkey LoLvl - der aktuelle Low-Level Spannungspegel erscheint im
Eingabefeld.
4 - Grundlegende Einrichtungsbeispiele
Einrichten einer Impulswellenform
107 TGF4000 Series Quick Start Guide
Geben Sie einen neuen Pegelwert mit der Nummerntastatur ein. Drücken Sie - . 6
Drücken Sie den Softkey V, um einen Low-Pegel von -600 mV zu bestigen.
Schalten Sie den Ausgang ein
Drücken Sie die Taste Output1, um den Kanal 1-Ausgang einzuschalten.
Die Ausgangstaste 1 leuchtet orange, um den Einschaltzustand anzuzeigen
4 - Grundlegende Einrichtungsbeispiele
Einrichten weiterer Ausgangsoptionen
108 TGF4000 Series Quick Start Guide
Einrichten weiterer Ausgangsoptionen
Voraussetzung
In den früheren Beispielen wurde gezeigt, wie das Ausgangsmenü verwendet wird, um den
Ausgangspegel einzustellen (Amplitude plus Offset oder High Pegel plus Low Pegel) und den
Ausgang ein- oder auszuschalten. Dieses Beispiel zeigt die Einstellung der Ausgangsphase,
Ausgangspolarität, Lastimpedanz und Spannungsautomatik.
Startbedingungen
Bevor Sie beginnen, sollten Sie das Gerät auf Standardwerte zurücksetzen, wie in Abschnitt
Anfangsbedingungen
Öffnen Sie das Ausgangsmenü
Drücken Sie den Softkey Output
4 - Grundlegende Einrichtungsbeispiele
Einrichten weiterer Ausgangsoptionen
109 TGF4000 Series Quick Start Guide
Ändern der Ausgangsphase
Standardmäßig ist der Softkey Phase ausgewählt.
Geben Sie eine Phase von - 4 5 Grad ein.
Der eingestellte Phasenwinkel ist der Punkt in der Wellenformperiode, der mit dem Sync oder der
Triggerflanke übereinstimmt, d. h. es ist der Punkt in der Periode, an der die Wellenform beginnt.
Daher eilt eine negative Phaseneinstellung voraus, während eine positive Phaseneinstellung die
Wellenform in Bezug auf den Sync oder Trigger verzögert. Dies wird auch im Grafikbereich
angezeigt.
Der Softkey Aligndient zur Neuausrichtung der Phase bei Frequenzänderungen.
Ändern der Ausgangspolarität
Drücken Sie den Softkey Type, um die Ausgangspolarität zu invertieren.
Beachten Sie, dass wiederholtes Drücken der Taste Type einen Wechsel zwischen normal und
invertiert bewirkt.
4 - Grundlegende Einrichtungsbeispiele
Einrichten weiterer Ausgangsoptionen
110 TGF4000 Series Quick Start Guide
Ändern der Lastimpedanz
Drücken Sie den Softkey Load
Drücken Sie den Softkey Load erneut, um die Lastimpedanz auf High-z(hohe Impedanz)
zu ändern.
Beachten Sie, dass die Standard-Lastimpedanz 50 Ohm beträgt, jedoch auf eine beliebige Impedanz
zwischen 50 und 10.000 Ohm eingestellt werden kann. Die Pegel werden auf der Grundlage dieser
Impedanz berechnet. Beachten Sie, dass wiederholtes Drücken der Taste Load einen Wechsel
zwischen einem numerischen Wert und High-zbewirkt. Beachten Sie, dass sich der angezeigte
Amplitudenwert auf 2 Volt pk-pk erhöht.
Drücken Sie den Softkey Load, um die Lastimpedanz wieder auf 50 Ohm zu setzen.
4 - Grundlegende Einrichtungsbeispiele
Einrichten weiterer Ausgangsoptionen
111 TGF4000 Series Quick Start Guide
Bereichsumschaltung
Drücken Sie den Softkey Range
Drücken Sie den Softkey Range
erneut, um den Bereich von Auto auf Hold zu ändern.
Die Spannungsautomatik verläuft in Dämpfungsschritten von jeweils 6dB (d. h. Range
Maximalwerte von 10 Vpp, 5 Vpp, 2,5 Vpp etc. an 50 ), wobei die Feineinstellung der Amplitude
auf 6dB begrenzt ist, um die Güte der Wellenform zu erhalten. Durch Wahl von Hold wird die
automatische Bereichswahl deaktiviert. Die Dämpfungseinstellung ist fixiert und die Feineinstellung
der Amplitude nicht länger begrenzt.
Wenn Rangeauf Auto eingestellt ist, schalten Amplitude und Abschwächer automatisch, sodass
eine optimale Leistung gegeben ist. Bei Bereichseinstellung Holdwird für alle
Amplitudeneinstellungen eine feste Dämpfungseinstellung verwendet.
5 - Erkunden der Generator-Funktionen
Einrichten einer Arbiträr-Wellenform
112 TGF4000 Series Quick Start Guide
5. ERKUNDEN DER GENERATOR-FUNKTIONEN
In den folgenden Beispielen werden nur die Parametereinstellungen beschrieben, zusammen mit
den dazugehörigen Tastenbezeichnungen. Es werden zudem die resultierenden
Ausgangswellenformen gezeigt (zusammen mit der Sync- oder Triggerwellenform, falls relevant).
Ausgangsamplitude und Offset-Einstellungen sind nur Beispiele und müssen nicht beachtet
werden.
Einrichten einer Arbiträr-Wellenform
Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück.
MENÜ
TASTENBEZEICHNUNG
Wellenform
Waves
MENÜ
Softkey-Bezeichnung
Arbiträr-Wellenform
Arb
Parameter
Softkey-Bezeichnung
Einstellung
Kurvenauswahl
Waves
Sinc
Frequenz
-
10 kHz
Amplitude
Ampl
2V
Offset
Offset
5mV
Parameter
TASTENBEZEICHNUNG
Einstellung
Ausgangsstatus
Output1
On
5 - Erkunden der Generator-Funktionen
Einrichten einer AM-modulierten Sinuskurve
113 TGF4000 Series Quick Start Guide
Einrichten einer AM-modulierten Sinuskurve
Setzen Sie das Get zunächst auf die Standardeinstellungen zurück.
Parameter
Softkey-Bezeichnung
Einstellung
Frequenz
-
10MHz
MENÜ
TASTENBEZEICHNUNG
Modulation
Mod
Parameter
Softkey-Bezeichnung
Einstellung
Frequenz
-
100kHz
Tiefe
Depth
100%
Quelle
Source
Internal
Form
Shape
Sinus
Modulationstatus
On/Off
On
Parameter
TASTENBEZEICHNUNG
Einstellung
Ausgangsstatus
Output1
On
Modulation auf AM-SC ändern
Parameter
Softkey-Bezeichnung
Einstellung
Typ
Type > AM
AM-SC
5 - Erkunden der Generator-Funktionen
PRBS
114 TGF4000 Series Quick Start Guide
PRBS
Setzen Sie das Get zunächst auf die Standardeinstellungen zurück.
MENÜ
TASTENBEZEICHNUNG
Wellenform
Waves
MENÜ
Softkey-Bezeichnung
Einstellung
Rauschen/PRBS
Noise
Quelle
Source
PBRS
Parameter
Softkey-Bezeichnung
Einstellung
Bit-Rate
BitRate
1Mbps
Amplitude
Ampl
3.3V
Offset
Offset
1.65V
PRBS Typ
Type
PN7
Parameter
TASTENBEZEICHNUNG
Einstellung
Ausgangsstatus
Output1
On
5 - Erkunden der Generator-Funktionen
Frequenzmodulation einer Sinuskurve
115 TGF4000 Series Quick Start Guide
Frequenzmodulation einer Sinuskurve
Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück.
MENÜ
TASTENBEZEICHNUNG
Modulation
Mod
Parameter
Softkey-Bezeichnung
Einstellung
Modulationsstatus
On/Off
On
Modulationstyp
Type
FM
Modulationsfrequenz
-
1kHz
Abweichung
Deviatn
9kHz
MENÜ
TASTENBEZEICHNUNG
Sinus
Params
Parameter
Softkey-Bezeichnung
Einstellung
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
TASTENBEZEICHNUNG
Einstellung
Ausgangsstatus
Output1
On
5 - Erkunden der Generator-Funktionen
Pulsweitenmodulierte Wellenform (PWM)
116 TGF4000 Series Quick Start Guide
Pulsweitenmodulierte Wellenform (PWM)
Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück.
MENÜ
TASTENBEZEICHNUNG
Wellenform
Waves
Parameter
Softkey-Bezeichnung
Puls
Pulse
MENÜ
TASTENBEZEICHNUNG
Modulation
Mod
Parameter
Softkey-Bezeichnung
Einstellung
Modulationsstatus
On/Off
On
Modulationstyp
Type
PWM
Modulationsfrequenz
-
1kHz
Abweichung
Dev %
40%
MENÜ
TASTENBEZEICHNUNG
Pulse
Params
Parameter
Softkey-Bezeichnung
Einstellung
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
TASTENBEZEICHNUNG
Einstellung
Ausgangsstatus
Output1
On
5 - Erkunden der Generator-Funktionen
Amplitudenumtastung (ASK)
117 TGF4000 Series Quick Start Guide
Amplitudenumtastung (ASK)
Setzen Sie das Get zunächst auf die Standardeinstellungen zurück.
MENÜ
TASTENBEZEICHNUNG
Modulation
Mod
Parameter
Softkey-Bezeichnung
Einstellung
Modulationsstatus
On/Off
On
Modulationstyp
Type
ASK
Modulationsquelle
Source
Internal
Hop-Amplitude
HpAmpl
100mV
Umtastrate
Rate
1kHz
Hop-Polarität
HopPol
Positive
MENÜ
TASTENBEZEICHNUNG
Sinus
Params
Parameter
Softkey-Bezeichnung
Einstellung
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
TASTENBEZEICHNUNG
Einstellung
Ausgangsstatus
Output1
On
Beachten Sie, dass die zweite Messkurve das Ausgangssignal der Hauptbuchse von Ausgang 2
darstellt, wenn Kanal 2 synchronisiert ist und dem modulierenden Wellenformsignal folgt.
5 - Erkunden der Generator-Funktionen
Frequenz-Sweep einer Sinuswelle
118 TGF4000 Series Quick Start Guide
Frequenz-Sweep einer Sinuswelle
Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück.
MENÜ
TASTENBEZEICHNUNG
Sweep
Sweep
Parameter
Softkey-Bezeichnung
Einstellung
Sweep-Status
On/Off
On
Stoppfrequenz
Freq > Stop
100kHz
MENÜ
TASTENBEZEICHNUNG
Sinus
Params
Parameter
Softkey-Bezeichnung
Einstellung
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
TASTENBEZEICHNUNG
Einstellung
Ausgangsstatus
Output1
On
5 - Erkunden der Generator-Funktionen
Generieren eines getriggerten Burst-Signals
119 TGF4000 Series Quick Start Guide
Generieren eines getriggerten Burst-Signals
Setzen Sie das Get zunächst auf die Standardeinstellungen zurück.
Parameter
Softkey-Bezeichnung
Einstellung
Frequenz
-
6MHz
MENÜ
TASTENBEZEICHNUNG
Burst
Burst
Parameter
Softkey-Bezeichnung
Einstellung
Burst-Status
On/Off
On
Burst-Count
Count
3
Triggerquelle
SetTrg > Source > Int
Internal Trigger
Triggerperiode
SetTrg > Period
5ms
MENÜ
TASTENBEZEICHNUNG
Sinus
Params
Parameter
Softkey-Bezeichnung
Einstellung
Amplitude
Ampl
1.0V
Offset
Offset
0.0V
Parameter
TASTENBEZEICHNUNG
Einstellung
Ausgangsstatus
Output1
On
Beachten Sie, dass die zweite Messkurve das Ausgangssignal der Hauptbuchse von Ausgang 2
darstellt, wenn Kanal 2 synchronisiert ist und dem modulierenden Trigger-Eingangssignal folgt.
5 - Erkunden der Generator-Funktionen
Frequenzkopplung beider Kanäle
120 TGF4000 Series Quick Start Guide
Frequenzkopplung beider Kanäle
Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück.
MENÜ
TASTENBEZEICHNUNG
Utility
Utility
Parameter
Softkey-Bezeichnung
Einstellung
Frequenzen
Dual Ch >Freq
Coupled
Parameter
TASTENBEZEICHNUNG
Einstellung
Ausgangsstatus
Output1
On
Ausgangsstatus
Output2
On
MENÜ
TASTENBEZEICHNUNG
Sinus
Params
Parameter
Softkey-Bezeichnung
Einstellung
Phasenverschiebung
Output > Phase > Done
90 Degrees
Frequenz
Freq
1MHz
Beachten Sie, dass, wenn Kanal 1 auf 1MHz eingestellt wird, Kanal 2 ebenfalls auf 1MHz eingestellt
ist. Die 90-Grad-Phasenverschiebung zwischen den Kanälen ist zu erkennen.
5 - Erkunden der Generator-Funktionen
Frequenzzähler
121 TGF4000 Series Quick Start Guide
Frequenzzähler
Setzen Sie das Get zunächst auf die Standardeinstellungen zurück.
MENÜ
TASTENBEZEICHNUNG
Utility
Utility
Parameter
Softkey-Bezeichnung
Einstellung
Zähler
Instr > FrCntr > On/Off
Counter Enabled
Quelle
Source
TRIG IN - DC-Coupled
Typ
Freq
Frequency
Messung
Count
Der Editierbereich zeigt die aktuelle Messung. Wenn am ausgewählten Eingang kein Eingangssignal
anliegt, zeigt der Zähler No signalan, bis ein entsprechendes Eingangssignal anliegt.
Wenn ein ausreichendes Eingangssignal anliegt, misst der Zähler ständig und zeigt den Messwert
im Editierbereich an.
6 - Wartung
Reinigung
122 TGF4000 Series Quick Start Guide
6. WARTUNG
Die Hersteller oder ihre Vertretungen bieten einen Reparaturdienst für fehlerhafte Geräte an. Falls
Anwender Wartungsarbeiten selbst durchführen möchten, sollten sie nur geschultes Personal damit
beauftragen. Für diese Arbeiten sollte das Servicehandbuch zu Hilfe genommen werden, das direkt
beim Hersteller der Geräte oder dessen Vertretungen bezogen werden kann.
Reinigung
Verwenden Sie zur Reinigung des Geräts ein leicht mit Wasser oder einem milden Reinigungsmittel
angefeuchtetes Tuch.
ACHTUNG! ZUR VERMEIDUNG VON STROMSCHLÄGEN ODER BESCHÄDIGUNGEN DES GERÄTS DARF KEIN
WASSER IN DAS GEHÄUSE GELANGEN. DAS GERÄT NICHT MIT
LÖSUNGSMITTELN REINIGEN, UM SCHÄDEN AM GEHÄUSE ZU VERMEIDEN.
0 - Indice
123 TGF4000 Series Quick Start Guide
INDICE
 Sicurezza 124
Simboli 125
 Principi di funzionamento 126
Layout del pannello anteriore 126
Layout del pannello posteriore 127
Layout dello schermo 128
 Per iniziare 129
Condizioni iniziali 129
 Esempi di configurazione di base 130
Configurazione di un segnale a onda sinusoidale 130
Configurazione di un segnale di clock di onda quadrata 134
Configurazione di una forma donda a impulso 140
Configurazione di altre opzioni di uscita 148
 Analisi delle funzionalità del generatore 152
Configurazione di un segnale donda arbitrario 152
Configurazione di una forma donda sinusoidale modulata AM 153
PRBS 154
Modulazione di frequenza di una forma donda sinusoidale 154
Forma donda modulata ad ampiezza di impulso (PWM) 156
Amplitude Shift Keying (ASK) (Modulazione a spostamento di ampiezza) 157
Frequenza di scansione di unonda sinusoidale 158
Generazione di un burst triggered 158
Accoppiamento della frequenza di entrambi i canali 160
Contatore di frequenza 161
 Manutenzione 162
Pulizia 162
Le ultime revisioni di questo manuale, driver di dispositivo e strumenti software possono essere
scaricati da: www.aimtti.com
1 - Sicurezza
124 TGF4000 Series Quick Start Guide
1. SICUREZZA
Il generatore è uno strumento di sicurezza classe I in base alla classificazione IEC ed è stato studiato
per corrispondere ai requisiti EN61010-1 (requisiti sulla sicurezza di attrezzature elettriche di
misurazione, controllo e uso in laboratorio). Euno strumento di installazione categoria II destinato
al funzionamento attraverso alimentazione a fase singola normale.
Lo strumento è stato testato in conformità a EN61010-1 ed è stato fornito in condizioni di sicurezza.
Il presente manuale di istruzioni comprende alcune informazioni e avvertenze che devono essere
seguite dallutente al fine di garantire un funzionamento sicuro e la conservazione dello strumento
in condizione di sicurezza.
Lo strumento è stato studiato per luso allinterno in ambiente con grado di inquinamento 2 con
range di temperatura da 5 °C a 40 °C, 20%-80% UR (senza formazione di condensa). Può essere
occasionalmente esposto a temperature comprese fra +5 °C e 10 °C senza compromissione della
sicurezza. Non utilizzare in presenza di condensa.
L’uso dello strumento in modo non specificato dalle presenti istruzioni può compromettere la
protezione di sicurezza fornita. Non utilizzare lo strumento al di fuori del tensione o delle condizioni
ambientali indicate.
AVVERTENZA! LO STRUMENTO DEVE PROVVISTO DI MESSA A TERRA
Qualsiasi interruzione del conduttore di terra allinterno o allesterno dello strumento rende
pericoloso luso dello strumento. Uninterruzione intenzionale è vietata. Non compromettere la
sicurezza mediante luso di una prolunga senza conduttore di protezione.
Quando lo strumento è collegato allalimentazione, i terminali possono essere scoperti e lapertura
delle coperture o la rimozione di parti (eccetto quelle accessibili con le mani) possono esporre parti
esposte. Prima dellapertura per regolazioni, sostituzioni, manutenzione o riparazione, lo
strumento deve essere scollegato da tutte le sorgenti di alimentazione.
Qualsiasi regolazione, manutenzione o riparazione dello strumento aperto in tensione deve essere
evitata e se inevitabile, deve essere effettuata esclusivamente da personale competente,
consapevole del possibile pericolo.
In caso di evidente difetto dello strumento, danno meccanico, umidità eccessiva o corrosione
chimica, la protezione di sicurezza potrebbe essere compromessa e il dispositivo non dovrà più
essere utilizzato e sottoposto a controllo e riparazione.
Accertare di utilizzare solo i fusibili della potenza e del tipo prescritti per eventuali sostituzioni.
L’uso di fusibili improvvisati e la cortocircuitazione del portafusibili è vietato.
Lo strumento utilizza una pila a bottone al litio per alimentare la memoria di sicurezza non volatile;
la pila dura in media 5 anni. Qualora sia necessario sostituirla, utilizzare una pila del tipo corretto,
ovvero una pila a bottone da 3 V Li/Mn0
2
da 20 mm tipo 2032. Smaltire le pile usate osservando le
disposizioni vigenti; non aprire, tagliare, bruciare od esporre la pila a temperature superiori ai 60°C
e non sottoporla a ricarica.
Non bagnare lo strumento durante la pulizia e utilizzare esclusivamente un panno asciutto per
pulire lo schermo LCD.
1 - Sicurezza
Simboli
125 TGF4000 Series Quick Start Guide
Simboli
I seguenti simboli vengono utilizzati sullo strumento e nel manuale:
Attenzione
- vedere i documenti allegati. Luso errato può
danneggiare lo strumento
terminale collegato al fondo dello chassis
alimentazione principale OFF.
l
alimentazione principale ON.
corrente alternata.
2 - Principi di funzionamento
Layout del pannello anteriore
126 TGF4000 Series Quick Start Guide
2. PRINCIPI DI FUNZIONAMENTO
Layout del pannello anteriore
Rif
Descrizione breve
Funzione
1
Interruttore di alimentazione
Accende e spegne lo strumento. Nota di sicurezza: Per scollegare
completamente l'alimentazione CA, staccare il cavo di
alimentazione dal retro dello strumento oppure scollegare dalla
presa di alimentazione CA. Assicurarsi che i punti di scollegamento
siano facilmente accessibili.
2
Soft-key:
Esegue la funzione mostrata nelletichetta del soft-key dellLCD
precedente.
3
Menu Waveform
Seleziona il tipo di forma donda principale (forma donda del
vettore) come attivo. (sinusoidi, quadre, a dente di sega, a impulso,
di rumore/PBRS e arbitrarie).
4
Menu Waveform
Modification
Apre il menu per limpostazione dei parametri di modulazione,
scansione e burst
5
Altri menu
Seleziona i menu per larchiviazione interna ed esterna dei file, le
utility degli strumenti e le condizioni di attivazione.
6
Socket principali
Socket di uscita principali. Il canale 2 può anche essere configurato
per la sincronizzazione con il canale di uscita 1 dal socket MAIN OUT
2.
7
Tasti uscita
Posizionare la MAIN OUT selezionata su On o Off.
8
Tasti cursore e rotellina
Utilizzato per cambiare i valori numerici dei parametri cifra per
cifra.
Utilizzato per selezionare le voci allinterno di alcuni menu.
9
Tastierino numerico
Utilizzato per inserire direttamente i valori numerici dei parametri.
10
Unità flash USB
Connettore dellhost USB per archiviazione nellunità flash USB.
2 - Principi di funzionamento
Layout del pannello posteriore
127 TGF4000 Series Quick Start Guide
Layout del pannello posteriore
Rif
Descrizione breve
Funzione
1
Ingresso modulazione
Ingresso esterno per modulazione delle forme donda principali.
2
Ingresso di
riferimento/Contatore di
frequenza accoppiato CA
Ingresso del clock di riferimento esterno da 10MHz e misurazione
della frequenza esterna accoppiata CA.
3
Uscita di riferimento
Uscita per un clock interno di riferimento da 10 MHz.
4
Ingresso di
generazione/Contatore di
frequenza accoppiato CC
Ingresso esterno per la generazione delle forme donda principali e
misurazione della frequenza esterna accoppiata CC
5
LAN
L'interfaccia LAN è stata progettata per soddisfare i requisiti LXI
Core 2011. Il protocollo TCP/IP Socket consente di attivare il
controllo a distanza tramite l'interfaccia LAN.
6
USB
Questa accetta cavi USB standard
7
GPIB (IEEE-488) - opzionale
L'indirizzo GPIB predefinito è 5.
2 - Principi di funzionamento
Layout dello schermo
128 TGF4000 Series Quick Start Guide
Layout dello schermo
Rif
Descrizione breve
Funzione
1
Indicatore di canale
Mostra il canale attualmente selezionato
2
Tipo di forma donda
principale
Mostra lattuale forma donda del vettore
3
Stato uscita
Indica se luscita principale è attivata o disattivata
4
Indicatore del clock esterno.
Mostra lo stato del clock esterno (se applicato)
5
Indicatore stato LAN
Mostra lo stato della connessione LAN (Ethernet).
6
Casella dei parametri
Mostra i parametri principali della forma donda.
7
Descrizione del menu
Mostra il menu di modifica attualmente selezionato.
8
Casella grafica
Mostra una rappresentazione grafica della forma donda
selezionata.
9
Casella di modifica
Mostra il parametro corrente che può essere modificato
10
Etichette soft-key
Mostra le funzioni correnti dei sei tasti seguenti.
3 - Per iniziare
Condizioni iniziali
129 TGF4000 Series Quick Start Guide
3. PER INIZIARE
Per familiarizzare lutente con alcune delle funzionalità di base dello strumento, in questa guida
vengono mostrati alcuni esempi di configurazione. Si consiglia a tutti gli utenti di applicare i primi
tre esempi:
Configurazione di un segnale a onda sinusoidale
Configurazione di un segnale di clock di onda quadrata
Configurazione di una forma donda a impulso
Configurazione di altre opzioni di uscita
Vengono forniti alcuni esempi di configurazione che presuppongono una maggiore familiarità con
lo strumento:
Configurazione di un segnale donda arbitrario
Configurazione di una forma donda sinusoidale modulata AM
PRBS (TGF4162 & TGF4242)
Modulazione di frequenza di una forma donda sinusoidale
Forma donda modulata ad ampiezza di impulso (PWM)
Amplitude Shift Keying (ASK) (Modulazione a spostamento di ampiezza)
Frequenza di scansione di unonda sinusoidale
Generazione di un burst triggered
Accoppiamento della frequenza di entrambi i canali
Contatore di frequenza
Condizioni iniziali
Prima di configurare lo strumento per uno degli esempi, deve essere ripristinato alle condizioni
predefinite.
A questo scopo, procedere come segue:
Premere il tasto fisso contrassegnato con Utility.
Premere il soft-key etichettato System
Premere il soft-key etichettato Default (lo schermo indica Restore Factory Default?)
Premere il soft-key etichettato Yes
La forma donda principale viene impostata Sinusoide (10kHz, 1V picco-picco) e vengono annullate
le modulazioni, le scansioni o il burst triggering o gating.
NOTA:
Lo strumento può essere impostato per conservare le impostazioni più recenti allo spegnimento e
ripristinarle allaccensione. Questa opzione viene impostata dal menu Utility e dal soft-key PwrOn.
Questa impostazione va perduta quando vengono ripristinate le condizioni predefinite dello
strumento, come sopra descritto.
4 - Esempi di configurazione di base
Configurazione di un segnale a onda sinusoidale
130 TGF4000 Series Quick Start Guide
4. ESEMPI DI CONFIGURAZIONE DI BASE
Configurazione di un segnale a onda sinusoidale
Requisito
Emettere un segnale continuo a onda sinusoidale con una frequenza di 40MHz e un’ampiezza di 6
volt picco-picco da MAIN OUT 1.
Condizioni di avvio
Prima dellavvio, ripristinare lo strumento alle impostazioni predefinite, come descritto nella
sezione Condizioni iniziali
Apertura del menu della forma donda - Sinusoidale
Premere il tasto fisso contrassegnato con Waves
Premere il soft-key etichettato Sine
4 - Esempi di configurazione di base
Configurazione di un segnale a onda sinusoidale
131 TGF4000 Series Quick Start Guide
Impostazione della frequenza
Il soft-key etichettato Freq viene evidenziato; la frequenza corrente viene mostrata nella casella di
modifica.
Tenere presente che premendo più volte il soft-key la funzione cambia tra Frequency e Period.
Utilizzare il tastierino numerico per inserire una nuova frequenza. Premere i numeri 4 0
Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di frequenza.
Premere il soft-key etichettato MHz per confermare una frequenza di 40MHz.
4 - Esempi di configurazione di base
Configurazione di un segnale a onda sinusoidale
132 TGF4000 Series Quick Start Guide
Impostazione dell’ampiezza
Premere il soft-key etichettato Ampl
Le successive pressioni del soft-key Ampl cambiano le etichette dei tasti Ampl e Offset in
HiLvl (livello alto) e LoLvl (livello basso) e viceversa.
Utilizzare il tastierino numerico per inserire una nuova ampiezza. Premere il numero 6
Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tensione.
4 - Esempi di configurazione di base
Configurazione di un segnale a onda sinusoidale
133 TGF4000 Series Quick Start Guide
Premere il soft-key etichettato Vpp per confermare unampiezza picco-picco di 6,0 volt.
Attivazione delluscita
Premere il tasto Output 1 per attivare l’uscita del canale 1.
Il tasto di uscita 1 si illumina in arancione per indicare lo stato attivo
4 - Esempi di configurazione di base
Configurazione di un segnale di clock di onda quadrata
134 TGF4000 Series Quick Start Guide
Configurazione di un segnale di clock di onda quadrata
Requisito
Emettere un segnale continuo di clock a onda quadrata con una frequenza di 20MHz, 50% ciclo di
funzionamento, un livello elevato di 3,3V e un livello basso di 0,0 volt da MAIN OUT 1.
Condizioni di avvio
Prima dellavvio, ripristinare lo strumento alle impostazioni predefinite, come descritto nella
Condizioni iniziali
Apertura del menu della forma donda - Quadrata
Premere il tasto fisso contrassegnato con Waves
Premere il soft-key etichettato Square.
4 - Esempi di configurazione di base
Configurazione di un segnale di clock di onda quadrata
135 TGF4000 Series Quick Start Guide
Impostazione della frequenza
Il soft-key etichettato Freq viene evidenziato; la frequenza corrente viene mostrata nella casella di
modifica.
Tenere presente che premendo più volte il soft-key la funzione cambia tra Frequency e Period.
Utilizzare il tastierino numerico per inserire una nuova frequenza. Premere i numeri 2 0.
Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di frequenza.
Premere il soft-key etichettato MHz per confermare una frequenza di 20MHz.
Tenere presente che la casella grafica cambia e mostra il tempo di salita ai fronti, che adesso è un
valore significativo.
4 - Esempi di configurazione di base
Configurazione di un segnale di clock di onda quadrata
136 TGF4000 Series Quick Start Guide
Conferma del ciclo di funzionamento
Premere il soft-key etichettato Duty; il ciclo di funzionamento corrente viene mostrato nella
casella di modifica.
Tenere presente che il ciclo di funzionamento è già impostato sul 50%, ma può essere modificato se
necessario.
Impostazione del livello alto e del livello basso
Premere il soft-key etichettato Ampl; l’etichetta del tasto cambia in HiLvl e la tensione del
livello alto corrente viene mostrata nella casella di modifica
Le successive pressioni del soft-key Ampl cambiano le etichette dei tasti Ampl e Offset in
HiLvl (livello alto) e LoLvl (livello basso) e viceversa.
4 - Esempi di configurazione di base
Configurazione di un segnale di clock di onda quadrata
137 TGF4000 Series Quick Start Guide
Quando il soft-key è etichettato HiLvl, la tensione del livello alto corrente viene mostrata nella
casella di modifica.
Utilizzare il tastierino numerico per inserire un nuovo livello. Premere i numeri 3 . 3
Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tensione.
Premere il soft-key etichettato V per confermare un livello alto di 3.3 volt.
4 - Esempi di configurazione di base
Configurazione di un segnale di clock di onda quadrata
138 TGF4000 Series Quick Start Guide
Premere il soft-key etichettato LoLvl; la tensione del livello basso corrente viene mostrata
nella casella di modifica.
Utilizzare il tastierino numerico per inserire un nuovo livello. Premere 0 .
Premere il soft-key etichettato V per confermare un livello basso di 0,0 volt.
Attivazione delluscita
Premere il tasto Output1 per attivare luscita del canale 1.
Il tasto di uscita 1 si illumina in arancione per indicare lo stato attivo
4 - Esempi di configurazione di base
Configurazione di un segnale di clock di onda quadrata
139 TGF4000 Series Quick Start Guide
Modifiche in tempo reale a un parametro numerico (es. frequenza)
I parametri numerici possono essere modificati utilizzando i tasti cursore e la rotellina in alternativa
al tastierino numerico.
Premere il tasto fisso contrassegnato con Waves
Premere il soft-key etichettato Square.
Premere il soft-key etichettato Freq; viene visualizzato il valore corrente della frequenza di
20,0MHz.
Premere i tasti fissi Cursor per spostare la selezione della modifica alla seconda cifra.
Usare la rotellina per modificare il valore; la frequenza viene immediatamente modificata.
NOTA:
Premere per attivare / disattivare la rotellina.
4 - Esempi di configurazione di base
Configurazione di una forma donda a impulso
140 TGF4000 Series Quick Start Guide
Configurazione di una forma donda a impulso
Requisito
Emettere un segnale continuo a impulsi con un periodo di 100ns, ampiezza di 30ns, durata dei
fronti di 20ns e un livello alto di 2,7V e un livello basso di -0,6 volt da MAIN OUT 1.
Condizioni di avvio
Prima dellavvio, ripristinare lo strumento alle impostazioni predefinite, come descritto nella
Condizioni iniziali
Apertura del menu della forma donda - Impulso
Premere il tasto fisso contrassegnato con Waves
Premere il soft-key etichettato Pulse.
4 - Esempi di configurazione di base
Configurazione di una forma donda a impulso
141 TGF4000 Series Quick Start Guide
Impostazione del periodo
Premere il soft-key etichettato PlsFrq in modo che cambi in PlsPer; viene visualizzato il
periodo corrente nella casella di modifica.
Tenere presente che premendo più volte il soft-key la funzione cambia tra Frequency e Period.
Utilizzare il tastierino numerico per inserire un nuovo periodo. Premere i numeri 1 0 0
Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tempo.
4 - Esempi di configurazione di base
Configurazione di una forma donda a impulso
142 TGF4000 Series Quick Start Guide
Premere il soft-key etichettato ns per confermare un periodo di 100ns.
Notare che la casella grafica cambia per mostrare una rappresentazione della durata dellimpulso e
dei fronti.
Impostazione dellampiezza dellimpulso
Premere il soft-key etichettato Duty
; l’etichetta cambia in Width e mostra la larghezza come
unità di tempo
4 - Esempi di configurazione di base
Configurazione di una forma donda a impulso
143 TGF4000 Series Quick Start Guide
Utilizzare il tastierino numerico per inserire una nuova ampiezza. Premere i numeri 3 0
Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tempo.
Premere il soft-key etichettato ns per confermare unampiezza di 30ns.
4 - Esempi di configurazione di base
Configurazione di una forma donda a impulso
144 TGF4000 Series Quick Start Guide
Impostazione della durata dei fronti dellimpulso
Premere il soft-key etichettato More
Premere il soft-key etichettato Edge
Usare i tasti cursore per selezionare la cifra che rappresenta unità di 10ns
Usare la rotellina per cambiare il valore in 20,0ns
Se si preferisce, il valore può essere inserito con il tastierino numerico.
Premere il soft-key etichettato Done.
4 - Esempi di configurazione di base
Configurazione di una forma donda a impulso
145 TGF4000 Series Quick Start Guide
Impostazione del livello alto e del livello basso
Premere il soft-key etichettato Ampl; l’etichetta del tasto cambia in HiLvl e la tensione del
livello alto corrente viene mostrata nella casella di modifica
Le successive pressioni del soft-key Ampl cambiano le etichette dei tasti Ampl e Offset in
HiLvl (livello alto) e LoLvl (livello basso) e viceversa.
Utilizzare il tastierino numerico per inserire un nuovo livello. Premere i numeri 2 . 7 .
Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tensione.
4 - Esempi di configurazione di base
Configurazione di una forma donda a impulso
146 TGF4000 Series Quick Start Guide
Premere il soft-key etichettato V per confermare un livello alto di 2,7 volt.
Premere il soft-key etichettato LoLvl; la tensione del livello basso corrente viene mostrata
nella casella di modifica.
4 - Esempi di configurazione di base
Configurazione di una forma donda a impulso
147 TGF4000 Series Quick Start Guide
Utilizzare il tastierino numerico per inserire un nuovo livello. Premere - . 6
Premere il soft-key etichettato V per confermare un livello basso di -600 mV.
Attivazione delluscita
Premere il tasto Output1 per attivare luscita del canale 1.
Il tasto di uscita 1 si illumina in arancione per indicare lo stato attivo
4 - Esempi di configurazione di base
Configurazione di altre opzioni di uscita
148 TGF4000 Series Quick Start Guide
Configurazione di altre opzioni di uscita
Requisito
Nei precedenti esempi di configurazione si è mostrato come usare il menu Output per impostare il
livello di uscita (ampiezza più offset o livello alto più livello basso) e attivare o disattivare l’uscita.
Questo esempio mostra limpostazione della fase di uscita, la polarità delluscita, limpedenza di
carico e la calibrazione automatica della tensione.
Condizioni di avvio
Prima dellavvio, ripristinare lo strumento alle impostazioni predefinite, come descritto nella
Condizioni iniziali
Aprire il menu Output
Premere il soft-key contrassegnato con Output
4 - Esempi di configurazione di base
Configurazione di altre opzioni di uscita
149 TGF4000 Series Quick Start Guide
Modifica della fase di uscita
Il soft-key etichettato come Phase viene selezionato per impostazione predefinita.
Inserire una fase di -45 gradi.
L’angolo di fase impostato è il punto nel periodo della forma donda che coincide con il fronte di
sincronizzazione o di generazione, ovvero il punto in cui si origina la forma donda. Quindi,
unimpostazione negativa della fase anticipa la forma donda in relazione alla sincronizzazione o alla
generazione, mentre unimpostazione positiva la ritarda. La forma donda nella casella grafica
cambia per riflettere questo fenomeno.
Il soft-key Align serve per riallineare la fase quando si modifica la frequenza.
Modifica della polarità dell’uscita
Premere il soft-key etichettato Type per invertire la polarità delluscita.
Le successive pressioni del tasto Type alternano tra polarità normale e invertita.
4 - Esempi di configurazione di base
Configurazione di altre opzioni di uscita
150 TGF4000 Series Quick Start Guide
Modifica dellimpedenza di carico
Premere il soft-key etichettato Load
Premere nuovamente il soft-key etichettato Load per cambiare limpedenza di carico in
High-z (alta impedenza).
L’impedenza di carico predefinita è 50 Ohm, ma può essere modificata in qualsiasi valore compreso
tra 50 e 10.000 Ohm. I livelli sono calcolati in base a questa impedenza.
Le successive pressioni del tasto Load alternano tra un valore numerico e High-z.
Tenere presente che il valore dellampiezza aumenta a 2 volt picco-picco.
Premere il soft-key etichettato Load per tornare allimpedenza di carico di 50 Omh.
4 - Esempi di configurazione di base
Configurazione di altre opzioni di uscita
151 TGF4000 Series Quick Start Guide
Modifica dellintervallo
Premere il soft-key etichettato Range
Premere nuovamente il soft-key etichettato Range per cambiare lintervallo da Auto a Hold.
La modalità Auto esegue automaticamente la calibrazione in incrementi dellattenuatore da 6dB
(ovvero, valori massimi dellintervallo di 10Vpp, 5Vpp, 2,5Vpp ecc. in 50), con lintervallo Vernier
dellampiezza limitato a 6dB per mantenere la qualità della forma donda.
Selezionando la modalità Hold viene disattivata la calibrazione automatica; limpostazione
dellattenuatore è fissa e lintervallo Vernier dellampiezza non è più limitato.
Con l'intervallo impostato su Auto l'ampiezza e gli attenuatori vengono cambiati automaticamente
e si ottiene la performance ottimale. Con lintervallo impostato su Hold viene utilizzata
unimpostazione fissa dellattenuatore per tutte le impostazioni dellampiezza.
5 - Analisi delle funzionalità del generatore
Configurazione di un segnale donda arbitrario
152 TGF4000 Series Quick Start Guide
5. ANALISI DELLE FUNZIONALITÀ DEL
GENERATORE
Nei seguenti esempi vengono descritte solo le impostazioni dei parametri, insieme ai nomi dei tasti
correlati. Vengono mostrate le forme donda delluscita risultanti, insieme alla forma donda di
sincronizzazione o generazione, dove pertinente. Le impostazioni di ampiezza delluscita e offset
sono solo esempi, che non è obbligatorio applicare.
Configurazione di un segnale donda arbitrario
Iniziare con lo strumento configurato con le impostazioni predefinite.
MENU
NOME DEL TASTO FISSO
Forma d'onda arbitraria
Waves
MENU
Nome del soft-key
Forma d'onda arbitraria
Arb
Parametro
Nome del soft-key
Impostazione
Selezione onda
Waves
Sinc
Frequenza
-
10kHz
Ampiezza
Ampl
2V
Offset
Offset
5mV
Parametro
NOME DEL TASTO FISSO
Impostazione
Stato uscita
Output1
On
5 - Analisi delle funzionalità del generatore
Configurazione di una forma donda sinusoidale modulata AM
153 TGF4000 Series Quick Start Guide
Configurazione di una forma donda sinusoidale modulata AM
Iniziare con lo strumento configurato con le impostazioni predefinite.
Parametro
Nome del soft-key
Impostazione
Frequenza
-
10MHz
MENU
NOME DEL TASTO FISSO
Modulazione
Mod
Parametro
Nome del soft-key
Impostazione
Frequenza
-
100kHz
Profondità
Depth
100%
Sorgente
Source
Internal
Forma
Shape
Sine
Stato modulazione
On/Off
On
Parametro
NOME DEL TASTO FISSO
Impostazione
Stato uscita
Output1
On
Cambio della modulazione in AM-SC
Parametro
Nome del soft-key
Impostazione
Tipo
Type > AM
AM-SC
5 - Analisi delle funzionalità del generatore
PRBS
154 TGF4000 Series Quick Start Guide
PRBS
Iniziare con lo strumento configurato con le impostazioni predefinite.
MENU
NOME DEL TASTO FISSO
Forma d'onda arbitraria
Waves
MENU
Nome del soft-key
Impostazione
Rumore/PRBS
Noise
Sorgente
Source
PBRS
Parametro
Nome del soft-key
Impostazione
Velocità di trasmissione
BitRate
1Mbps
Ampiezza
Ampl
3,3V
Offset
Offset
1,65V
Tipo PRBS
Type
PN7
Parametro
NOME DEL TASTO FISSO
Impostazione
Stato uscita
Output1
On
5 - Analisi delle funzionalità del generatore
Modulazione di frequenza di una forma donda sinusoidale
155 TGF4000 Series Quick Start Guide
Modulazione di frequenza di una forma donda sinusoidale
Iniziare con lo strumento configurato con le impostazioni predefinite.
MENU
NOME DEL TASTO FISSO
Modulazione
Mod
Parametro
Nome del soft-key
Impostazione
Stato modulazione
On/Off
On
Tipo modulazione
Type
FM
Frequenza di modulazione
-
1kHz
Deviazione
Deviatn
9kHz
MENU
NOME DEL TASTO FISSO
Sinusoide
Params
Parametro
Nome del soft-key
Impostazione
Ampiezza
Ampl
1,0V
Offset
Offset
0,0V
Parametro
NOME DEL TASTO FISSO
Impostazione
Stato uscita
Output1
On
5 - Analisi delle funzionalità del generatore
Forma donda modulata ad ampiezza di impulso (PWM)
156 TGF4000 Series Quick Start Guide
Forma donda modulata ad ampiezza di impulso (PWM)
Iniziare con lo strumento configurato con le impostazioni predefinite.
MENU
NOME DEL TASTO FISSO
Forma d'onda arbitraria
Waves
Parametro
Nome del soft-key
Impostazione
Impulso
Pulse
MENU
NOME DEL TASTO FISSO
Modulazione
Mod
Parametro
Nome del soft-key
Impostazione
Stato modulazione
On/Off
On
Tipo modulazione
Type
PWM
Frequenza di modulazione
-
1kHz
Deviazione
Dev %
40%
MENU
NOME DEL TASTO FISSO
Impulso
Params
Parametro
Nome del soft-key
Impostazione
Ampiezza
Ampl
1,0V
Offset
Offset
0,0V
Parametro
NOME DEL TASTO FISSO
Impostazione
Stato uscita
Output1
On
5 - Analisi delle funzionalità del generatore
Amplitude Shift Keying (ASK) (Modulazione a spostamento di ampiezza)
157 TGF4000 Series Quick Start Guide
Amplitude Shift Keying (ASK) (Modulazione a spostamento di
ampiezza)
Iniziare con lo strumento configurato con le impostazioni predefinite.
MENU
NOME DEL TASTO FISSO
Modulazione
Mod
Parametro
Nome del soft-key
Impostazione
Stato modulazione
On/Off
On
Tipo modulazione
Type
ASK
Sorgente di modulazione
Source
Internal
Ampiezza Hop
HpAmpl
100mV
Velocità di commutazione
Rate
1kHz
Polarità Hop
HopPol
Positive
MENU
NOME DEL TASTO FISSO
Sinusoide
Params
Parametro
Nome del soft-key
Impostazione
Ampiezza
Ampl
1,0V
Offset
Offset
0,0V
Parametro
NOME DEL TASTO FISSO
Impostazione
Stato uscita
Output1
On
Notare che la seconda traccia è luscita dal socket principale delluscita 2 quando il canale 2 è
sincronizzato, che segue il segnale di forma donda di modulazione.
5 - Analisi delle funzionalità del generatore
Frequenza di scansione di unonda sinusoidale
158 TGF4000 Series Quick Start Guide
Frequenza di scansione di unonda sinusoidale
Iniziare con lo strumento configurato con le impostazioni predefinite.
MENU
NOME DEL TASTO FISSO
Scansione
Sweep
Parametro
Nome del soft-key
Impostazione
Stato scansione
On/Off
On
Frequenza arresto
Freq > Stop
100kHz
MENU
NOME DEL TASTO FISSO
Sinusoide
Params
Parametro
Nome del soft-key
Impostazione
Ampiezza
Ampl
1,0V
Offset
Offset
0,0V
Parametro
NOME DEL TASTO FISSO
Impostazione
Stato uscita
Output1
On
5 - Analisi delle funzionalità del generatore
Generazione di un burst triggered
159 TGF4000 Series Quick Start Guide
Generazione di un burst triggered
Iniziare con lo strumento configurato con le impostazioni predefinite.
Parametro
Nome del soft-key
Impostazione
Frequenza
-
6MHz
MENU
NOME DEL TASTO FISSO
Burst
Burst
Parametro
Nome del soft-key
Impostazione
Stato burst
On/Off
On
Conteggio dei burst
Count
3
Origine trigger
SetTrg > Source > Int
Internal Trigger
Periodo trigger
SetTrg > Period
5ms
MENU
NOME DEL TASTO FISSO
Sinusoide
Params
Parametro
Nome del soft-key
Impostazione
Ampiezza
Ampl
1,0V
Offset
Offset
0,0V
Parametro
NOME DEL TASTO FISSO
Impostazione
Stato uscita
Output1
On
Notare che la seconda traccia è luscita dal socket principale delluscita 2 quando il canale 2 è
sincronizzato, che segue il segnale di forma di ingresso trigger.
5 - Analisi delle funzionalità del generatore
Accoppiamento della frequenza di entrambi i canali
160 TGF4000 Series Quick Start Guide
Accoppiamento della frequenza di entrambi i canali
Iniziare con lo strumento configurato con le impostazioni predefinite.
MENU
NOME DEL TASTO FISSO
Utility
Utility
Parametro
Nome del soft-key
Impostazione
Frequenze
Dual Ch > Freq
Coupled
Parametro
NOME DEL TASTO FISSO
Impostazione
Stato uscita
Output1
On
Stato uscita
Output2
On
MENU
NOME DEL TASTO FISSO
Sinusoide
Parmas
Parametro
Nome del soft-key
Impostazione
Spostamento di fase
Output > Phase > Done
90 gradi
Frequenza
Freq
1MHz
Se il canale 1 è impostato su 1MHz, anche il canale 2 è impostato su 1MHz.
Si può vedere lo spostamento di fase di 90 gradi tra i canali.
5 - Analisi delle funzionalità del generatore
Contatore di frequenza
161 TGF4000 Series Quick Start Guide
Contatore di frequenza
Iniziare con lo strumento configurato con le impostazioni predefinite.
MENU
NOME DEL TASTO FISSO
Utility
Utility
Parametro
Nome del soft-key
Impostazione
Contatore
Instr > FrCntr > On/Off
Counter Enabled
Sorgente
Source
TRIG IN - DC coupled
Tipo
Freq
Frequency
Misurazione
Count
La casella di modifica mostra la misurazione corrente. Se non è presente alcun segnale in entrata
allingresso selezionato, il contatore segna No signal’ finché non viene applicato un segnale in entrata
adeguato.
Quando viene applicato un segnale di ingresso adeguato, il contatore misura costantemente e visualizza
il valore nella casella di modifica.
6 - Manutenzione
Pulizia
162 TGF4000 Series Quick Start Guide
6. MANUTENZIONE
I Produttori o i loro agenti allestero faranno le riparazioni necessarie in caso di guasto. The
Manufacturers or their agents overseas will provide a repair service for any unit developing a fault. Nel
caso in cui i proprietari desiderano effettuare da sé i lavori di manutenzione, questi devono essere svolti
esclusivamente da personale qualificato in conformità con il manuale di servizio, che può essere
ottenuto direttamente dai produttori o dai propri agenti distaccati.
Pulizia
Se si deve pulire il dispositivo di alimentazione, usare uno strofinaccio appena bagnato con acqua o con
un detergente neutro. Pulire la finestrella di visualizzazione con un panno asciutto e morbido.
ATTENZIONE! PER EVITARE SCOSSE ELETTRICHE ED EVENTUALI DANNI AL DISPOSITIVO DI
ALIMENTAZIONE, NON PERMETTERE MAI ALLACQUA DI ENTRARE ALLINTERNO DELLALLOGGIAMENTO.
PER EVITARE DANNI ALLALLOGGIAMENTO E ALLA FINESTRELLA DI VISUALIZZAZIONE, NON PULIRE MAI
CON SOLVENTI.
0 - Índice
163 TGF4000 Series Quick Start Guide
ÍNDICE
 Seguridad......................................................................................................................................................... 164
Símbolos ............................................................................................................................................................... 165
 Principios de funcionamiento .......................................................................................................................... 166
Distribución del panel frontal ............................................................................................................................... 166
Distribución del panel posterior ........................................................................................................................... 167
Distribución de la pantalla .................................................................................................................................... 168
 Primeros pasos ................................................................................................................................................ 169
Estado inicial ......................................................................................................................................................... 169
 Ejemplos básicos de configuración ................................................................................................................... 170
Configurar una señal de onda senoidal ................................................................................................................ 170
Configurar una señal de reloj de onda cuadrada .................................................................................................. 174
Configurar una forma de onda de pulso ............................................................................................................... 180
Configurar más opciones de salida ....................................................................................................................... 188
 Repaso de las capacidades del generador ........................................................................................................ 192
Configurar una señal de onda arbitraria ............................................................................................................... 192
Configurar una forma de onda senoidal con modulación AM .............................................................................. 193
SBSA ..................................................................................................................................................................... 194
Modulación de frecuencia de una forma de onda de senoidal ............................................................................ 194
Forma de onda modulada por anchura de pulso (PWM) ..................................................................................... 196
Modulación por desplazamiento de amplitud (ASK) ............................................................................................ 196
Barrido de frecuencia de onda senoidal ............................................................................................................... 198
Generación de una ráfaga activada ...................................................................................................................... 199
Emparejamiento de la frecuencia de ambos canales ........................................................................................... 200
Contador de frecuencia ........................................................................................................................................ 201
 Mantenimiento ................................................................................................................................................ 202
Limpieza ............................................................................................................................................................... 202
 Specification .................................................................................................................................................... 203
Las últimas revisiones de este manual, controladores de dispositivos y herramientas de software se
pueden descargar desde: www.aimtti.com
1 - Seguridad
164 TGF4000 Series Quick Start Guide
1. SEGURIDAD
El presente generador constituye un instrumento que pertenece a la Clase de Seguridad I de la
clasificación CEI y ha sido diseñada para cumplir las prescripciones de la norma EN 610101
(Requisitos de seguridad de equipos eléctricos de medida, control y uso en laboratorio). Se trata de
un instrumento de la Categoría de Instalación II que se debe alimentar con un suministro
monofásico normal.
Este instrumento se ha sometido a pruebas con arreglo a la norma EN 610101 y se proporciona
en condiciones de funcionamiento seguro. El presente manual de instrucciones contiene
información y advertencias que el usuario debe seguir, con el fin de garantizar y perpetuar la
seguridad de funcionamiento.
Este instrumento ha sido diseñado para su uso en interiores, en entornos con una polución de
grado 2 y dentro de un intervalo de temperaturas comprendido entre 5 °C y 40 °C, con una
humedad relativa comprendida entre el 20 % y el 80 % (sin condensación). Se puede someter
ocasionalmente a temperaturas comprendidas entre +5 °C y 10 °C, sin que su seguridad se vea
reducida. No se debe utilizar cuando haya condensación.
El uso de este instrumento de forma distinta a la especificada en estas instrucciones puede afectar
a sus mecanismos de seguridad. No utilice el instrumento con tensiones ni en entornos que se
encuentren fuera del intervalo especificado.
¡ADVERTENCIA! ESTE INSTRUMENTO DEBE CONECTARSE A TIERRA
Cualquier interrupción del conductor de puesta a tierra, dentro o fuera del instrumento, hará que
este resulte peligroso. Está prohibida la interrupción intencionada. No se debe inhibir este
mecanismo de protección mediante un alargador que no tenga conductor de toma de tierra.
Cuando el instrumento esté conectado a la fuente de alimentación puede haber terminales con
tensión y es probable que, si se abre la carcasa o se retiran piezas a las que no sea posible acceder
manualmente en condiciones normales, queden al descubierto componentes con tensión. Es
necesario desconectar el instrumento de cualquier fuente de alimentación antes de abrirlo para
realizar tareas de ajuste, sustitución, mantenimiento o reparación.
Se debe evitar en la medida de lo posible la realización de cualquier tarea de ajuste,
mantenimiento o reparación del instrumento abierto con tensión y, si fuera inevitable, solo la
realizará una persona con la preparación suficiente y que conozca los peligros inherentes.Si el
instrumento resultara estar claramente defectuoso, hubiera sido sometido a un daño mecánico, a
humedad excesiva o a corrosión química, su protección de seguridad podría fallar, por lo que será
necesario dejar de utilizar el aparato y devolverlo para su comprobación y reparación.
Asegúrese de que, cuando sea necesaria una sustitución, se utilicen únicamente fusibles de la
corriente nominal y el tipo especificados. Está prohibido utilizar fusibles artesanales o cortocircuitar
los portafusibles.
Este instrumento usa una pila de botón de litio como bateríade respaldo para la memoria no
volátil; su duración típica es de 5 años. De ser necesario, sustitúyala solo por otra del tipo
correcto: una pila de botón de Li/MnO
2
de 3 V, tipo 2032 (20 mm). Las pilas agotadas deben
eliminarse de forma respetuosa y de acuerdo con la legislación local; no las abra, incinere ni
exponga a temperaturas por encima de 60 °C, y no intente recargarlas.
No se debe humedecer el instrumento durante la limpieza; en particular, la pantalla LCD se debe
limpiar únicamente con un paño suave y seco.
1 - Seguridad
Símbolos
165 TGF4000 Series Quick Start Guide
Símbolos
En el instrumento y en este manual se utilizan los siguientes símbolos:
Precaución
Remítase a los documentos adjuntos; un uso
incorrecto puede dañar el instrumento.
Terminal conectado a tierra.
Alimentación de red OFF (desconectada).
l
Alimentación de red ON (conectada).
Corriente alterna (CA).
2 - Principios de funcionamiento
Distribución del panel frontal
166 TGF4000 Series Quick Start Guide
2. PRINCIPIOS DE FUNCIONAMIENTO
Distribución del panel frontal
Ref.
Descripción breve
Función
1
Interruptor de alimentación
Enciende y apaga el instrumento. Nota de seguridad: Para
desconectarlo completamente de la corriente alterna,
desenchufe el cable de alimentación en la parte posterior del
instrumento o en el punto de la toma de red; asegúrese de
tener siempre accesible un medio de desconexión.
2
Teclas variables
Ejecutan la función indicada en la etiqueta situada en la pantalla
LCD sobre ella.
3
Menús de forma de onda
Seleccionan el tipo principal de forma de onda activa, es decir, la
forma de onda portadora (senoidal, cuadrada, rampa, pulso,
ruido/SBSA o arbitraria).
4
Menús de modificación de forma
de onda
Abre los menús de configuración de los parámetros de
modulación, barrido y ráfaga.
5
Otros menús
Selecciona los menús de almacenamiento interno y externo de
archivos, utilidades del instrumento y condiciones de activación.
6
Tomas principales
Tomas principales de salida. El canal 2 también puede
configurarse para transportar la sincronía del canal 1 desde su
toma MAIN OUT 2.
7
Teclas de salida
Activan o desactivan la salida principal MAIN OUT seleccionada.
8
Teclas de cursor y mando
giratorio.
Usadas para cambiar los valores de los parámetros numéricos
dígito a dígito.
Usadas para seleccionar elementos en algunos menús.
9
Teclado numérico
Usado para introducir valores numéricos directamente.
10
Almacenamiento USB
Conector host USB para unidad de memoria USB.
2 - Principios de funcionamiento
Distribución del panel posterior
167 TGF4000 Series Quick Start Guide
Distribución del panel posterior
Ref.
Descripción breve
Función
1
Entrada de modulación
Entrada externa para modulación de las formas de onda
principales.
2
Entrada de referencia/Contador de
frecuencia con acoplamiento de CA
Entrada para reloj de referencia a 10 MHz y medición externa
de frecuencia con acoplamiento de CA.
3
Salida de referencia
Salida para reloj de referencia interna a 10 MHz.
4
Entrada de activación/Contador de
frecuencia con acoplamiento de CC
Entrada externa para activar las formas de onda principales y
medición externa de frecuencia con acoplamiento de CC.
5
LAN (red de área local)
La interfaz LAN está diseñada para ser compatible con el
estándar LXI Core 2011. El control remoto a través de la interfaz
LAN es posible mediante el protocolo de sockets de TCP/IP.
6
USB
El puerto USB, que acepta un cable USB estándar.
7
GPIB (IEEE
488): opcional
La dirección GPIB por defecto es 5.
2 - Principios de funcionamiento
Distribución de la pantalla
168 TGF4000 Series Quick Start Guide
Distribución de la pantalla
Ref.
Descripción breve
Función
1
Indicador de canal
Muestra el canal actualmente seleccionado
2
Tipo de forma de onda principal
Muestra la forma de onda portadora actual
3
Estado de salida
Muestra si la salida principal está encendida o apagada.
4
Indicador de señal de reloj externa
Muestra el estado de la señal de reloj externa (si la hubiera).
5
Indicador del estado de la LAN
Muestra el estado de la conexión de área local (Ethernet).
6
Caja de parámetros
Muestra los principales parámetros de la forma de onda.
7
Descripción del me
Muestra el menú de edición actualmente seleccionado.
8
Caja de gráficas
Muestra una representación gráfica de la forma de onda
seleccionada.
9
Caja de edición
Muestra el parámetro actual que puede ser editado.
10
Etiquetas de teclas variables
Muestra las funciones actuales de las seis teclas situadas
debajo.
3 - Primeros pasos
Estado inicial
169 TGF4000 Series Quick Start Guide
3. PRIMEROS PASOS
A fin de que el usuario se familiarice con algunas de las funcionalidades básicas del instrumento,
esta guía presenta una serie de ejemplos de configuración.
Se recomienda a todos los usuarios proceder con los tres primeros ejemplos:
Configurar una señal de onda senoidal
Configurar una señal de reloj de onda cuadrada
Configurar una forma de onda de pulso
Configurar más opciones de salida
Se incluyen una serie de ejemplos de configuración adicionales que asumen cierta familiarización
con el instrumento.
Configurar una señal de onda arbitraria
Configurar una forma de onda senoidal con modulación AM
SBSA (TGF4162 & TGF4242)
Modulación de frecuencia de una forma de onda de senoidal
Forma de onda modulada por anchura de pulso (PWM)
Modulación por desplazamiento de amplitud (ASK)
Barrido de frecuencia de onda senoidal
Generación de una ráfaga activada
Emparejamiento de la frecuencia de ambos canales
Contador de frecuencia
Estado inicial
Antes de configurar el instrumento para cualquiera de los ejemplos, deberá restablecerse su estado
por defecto.
Para ello, lleve a cabo los siguientes pasos:
Pulse la tecla física Utility (utilidad).
Pulse la tecla variable System (sistema).
Pulse la tecla variable Default (por defecto): la pantalla mostrará Restore Factory
Default? (¿restaurar valores de fábrica por defecto?).
Pulse la tecla variable Yes (sí).
Estas acciones establecen el seno como forma de onda principal (10 kHz, 1 Vpp) y cancelan
cualquier modulación, barrido o activación/conmutación de ráfaga.
NOTA:
Es posible configurar el instrumento para que recuerde sus últimos valores al apagarlo y los
restaure al encenderse. Esto se realiza desde el menú Utility, con la tecla variable PwrOn
(encendido). Esta configuración se perderá cuando se restablezca el estado por defecto del
instrumento según lo anteriormente descrito.
4 - Ejemplos básicos de configuración
Configurar una señal de onda senoidal
170 TGF4000 Series Quick Start Guide
4. EJEMPLOS BÁSICOS DE CONFIGURACIÓN
Configurar una señal de onda senoidal
Requisito
Emitir una señal de onda senoidal con una frecuencia de 40 MHz y amplitud de 6 Vpp en la salida
MAIN OUT 1.
Estado inicial
Antes de empezar, restaure los valores por defecto del instrumento según lo descrito en la
secciónEstado inicial
Abrir menú de forma de onda senoidal
Pulse la tecla física Waves
Pulse la tecla variable Sine
4 - Ejemplos básicos de configuración
Configurar una señal de onda senoidal
171 TGF4000 Series Quick Start Guide
Configurar la frecuencia
Se resaltará la tecla variable Freq: en la caja de edición se mostrará la frecuencia actual.
Tenga en cuenta que, al pulsar repetidamente esta tecla variable, su función alterna entre
Frequency y Period.
Use el teclado numérico para introducir una nueva frecuencia. Pulse los números 4 0
Advierta cómo, tan pronto como los introduce, las teclas variables pasan a mostrar las unidades de
frecuencia.
Pulse la tecla variable MHz para confirmar una frecuencia de 40 MHz.
4 - Ejemplos básicos de configuración
Configurar una señal de onda senoidal
172 TGF4000 Series Quick Start Guide
Configurar la amplitud
Pulse la tecla variable Ampl.
Advierta cómo las sucesivas pulsaciones de la tecla variable Ampl alternan las etiquetas Ampl y
Offset (desplazamiento) con HiLvl (nivel alto) y LoLvl (nivel bajo) y viceversa.
Use el teclado numérico para introducir una nueva amplitud. Pulse el número 6
Advierta cómo, tan pronto como lo introduce, las teclas variables pasan a mostrar las unidades de
tensión.
4 - Ejemplos básicos de configuración
Configurar una señal de onda senoidal
173 TGF4000 Series Quick Start Guide
Pulse la tecla variable Vpp para confirmar una amplitud pico a pico de 6,0 voltios.
Encender la salida
Pulse la tecla Output 1 para encender la salida del canal 1.
La tecla de salida 1 se ilumina en naranja para indicar el estado de encendido
4 - Ejemplos básicos de configuración
Configurar una señal de reloj de onda cuadrada
174 TGF4000 Series Quick Start Guide
Configurar una señal de reloj de onda cuadrada
Requisito
Emitir una señal de reloj de onda cuadrada continua con una frecuencia de 20 MHz, ciclo de
trabajo del 50 %, nivel alto de 3,3 V y nivel bajo de 0,0 V desde MAIN OUT 1.
Estado inicial
Antes de empezar, restaure los valores por defecto del instrumento según lo descrito en la sección
Estado inicial
Abrir menú de forma de onda cuadrada
Pulse la tecla física Waves
Pulse la tecla variable Square
4 - Ejemplos básicos de configuración
Configurar una señal de reloj de onda cuadrada
175 TGF4000 Series Quick Start Guide
Configurar la frecuencia
Se resaltará la tecla variable Freq: en la caja de edición se mostrará la frecuencia actual.
Tenga en cuenta que, al pulsar repetidamente esta tecla variable, su función alterna entre
Frequency y Period.
Use el teclado numérico para introducir una nueva frecuencia. Pulse los números 2 0
Advierta cómo, tan pronto como los introduce, las teclas variables pasan a mostrar las unidades de
frecuencia.
Pulse la tecla variable MHz para confirmar una frecuencia de 20 MHz.
Advierta cómo la caja de gráficas cambia para mostrar el tiempo de ascenso en los flancos, por ser
ahora un aspecto significativo.
4 - Ejemplos básicos de configuración
Configurar una señal de reloj de onda cuadrada
176 TGF4000 Series Quick Start Guide
Confirmar el ciclo de trabajo
Pulse la tecla variable Duty (ciclo de trabajo): en la caja de edición se mostrará el ciclo de
trabajo actual.
Observe que el ciclo de trabajo ya está configurado al 50 %, pero puede cambiarse desde aquí si lo
desea.
Configurar el nivel alto y bajo
Pulse la tecla variable Ampl; su etiqueta cambia a HiLvl y en la caja de edición se muestra
la tensión actual del nivel alto.
Advierta cómo las sucesivas pulsaciones de la tecla variable Ampl alternan las etiquetas Ampl y
Offset (desplazamiento) con HiLvl (nivel alto) y LoLvl (nivel bajo) y viceversa.
4 - Ejemplos básicos de configuración
Configurar una señal de reloj de onda cuadrada
177 TGF4000 Series Quick Start Guide
Cuando la tecla variable indica HiLvl, en la caja de edición se muestra la tensión actual del nivel
alto.
Use el teclado numérico para introducir un nivel distinto. Pulse los números 3 . 3
Advierta cómo, tan pronto como lo introduce, las teclas variables pasan a mostrar las unidades de
tensión.
Pulse la tecla variable V para confirmar un nivel alto de 3,3 voltios.
4 - Ejemplos básicos de configuración
Configurar una señal de reloj de onda cuadrada
178 TGF4000 Series Quick Start Guide
Pulse la tecla variable LoLvl - en la caja de edición se muestra la tensión actual del nivel
bajo.
Use el teclado numérico para introducir un nivel distinto. Pulse 0
Pulse la tecla variable V para confirmar un nivel bajo de 0,0 voltios.
Encender la salida
Pulse la tecla Output1 para encender la salida del canal 1.
La tecla de salida 1 se ilumina en naranja para indicar el estado de encendido.
4 - Ejemplos básicos de configuración
Configurar una señal de reloj de onda cuadrada
179 TGF4000 Series Quick Start Guide
Cambio de cualquier parámetro numérico en tiempo real (p. ej. la frecuencia)
Los parámetros numéricos pueden cambiarse usando las teclas de cursor y el mando giratorio
como alternativa al teclado numérico.
Pulse la tecla física Waves
Pulse la tecla variable Square
Pulse la tecla variable Freq: se mostrará la frecuencia actual de 20,0 MHz.
Pulse las teclas físicas del cursor para mover el foco de edición al segundo dígito.
Use el mando giratorio para modificar el valor: la frecuencia se cambia de inmediato.
NOTA:
Presione para activar / desactivar la rueda de giro.
4 - Ejemplos básicos de configuración
Configurar una forma de onda de pulso
180 TGF4000 Series Quick Start Guide
Configurar una forma de onda de pulso
Requisito
Emitir una señal de pulso continua con un periodo de 100 ns, anchura de pulso de 30 ns, tiempos
de flanco de 20 ns, nivel alto de 2,7 V y nivel bajo de -0,6 V desde MAIN OUT 1.
Estado inicial
Antes de empezar, restaure los valores por defecto del instrumento según lo descrito en la sección
Estado inicial
Abrir menú de forma de onda de pulso
Pulse la tecla física Waves
Pulse la tecla variable Pulse.
4 - Ejemplos básicos de configuración
Configurar una forma de onda de pulso
181 TGF4000 Series Quick Start Guide
Configurar el periodo
Pulse la tecla variable PlsFreq (frecuencia de pulso) de modo que cambie a PlsPer
(periodo de pulso): la caja de edición pasará a mostrar el periodo actual.
Tenga en cuenta que, al pulsar repetidamente esta tecla, su función alterna entre Frequency y
Period.
Use el teclado numérico para introducir un periodo distinto. Pulse los números 1 0 0
Advierta cómo, tan pronto como los introduce, las teclas variables pasan a mostrar las unidades de
tiempo.
4 - Ejemplos básicos de configuración
Configurar una forma de onda de pulso
182 TGF4000 Series Quick Start Guide
Pulse la tecla variable ns para confirmar un periodo de 100 ns.
Observemo la caja de gráficas pasa a mostrar una representación de los tiempos de pulso y
flanco.
Configurar la anchura del pulso
Pulse la tecla variable Duty (ciclo de trabajo); su etiqueta cambia a Width (anchura) y se
muestra la anchura en unidades de tiempo.
4 - Ejemplos básicos de configuración
Configurar una forma de onda de pulso
183 TGF4000 Series Quick Start Guide
Use el teclado numérico para introducir una nueva anchura. Pulse los números 3 0
Advierta cómo, tan pronto como los introduce, las teclas variables pasan a mostrar las unidades de
tiempo.
Pulse la tecla variable ns para confirmar una anchura de 30 ns.
4 - Ejemplos básicos de configuración
Configurar una forma de onda de pulso
184 TGF4000 Series Quick Start Guide
Configurar los tiempos de flanco del pulso
Pulse la tecla variable More
Pulse la tecla variable Edge
Use las teclas de cursor para seleccionar el valor que representa las unidades de 10 ns.
Use el mando giratorio para cambiar el valor a 20,0 ns.
Pulse la tecla variable Done
.
Observe cómo el valor podría haberse introducido utilizando el teclado numérico si así lo desea.
4 - Ejemplos básicos de configuración
Configurar una forma de onda de pulso
185 TGF4000 Series Quick Start Guide
Configurar el nivel alto y bajo
Pulse la tecla variable Ampl; su etiqueta cambia a HiLvl y en la caja de edición se muestra
la tensión actual del nivel alto.
Advierta cómo las sucesivas pulsaciones de la tecla variable Ampl alternan las etiquetas Ampl y
Offset (desplazamiento) con HiLvl (nivel alto) y LoLvl (nivel bajo) y viceversa.
Use el teclado numérico para introducir un nivel distinto. Pulse los números 2 . 7 .
Advierta cómo, tan pronto como lo introduce, las teclas variables pasan a mostrar las unidades de
tensión.
4 - Ejemplos básicos de configuración
Configurar una forma de onda de pulso
186 TGF4000 Series Quick Start Guide
Pulse la tecla variable V para confirmar un nivel alto de 2,7 voltios.
Pulse la tecla variable LoLvl: en la caja de edición se muestra la tensión actual del nivel bajo.
4 - Ejemplos básicos de configuración
Configurar una forma de onda de pulso
187 TGF4000 Series Quick Start Guide
Use el teclado numérico para introducir un nivel distinto. Pulse - . 6
Pulse la tecla variable V para confirmar un nivel bajo de -600 mV.
Encender la salida
Pulse la tecla Output1 para encender la salida del canal 1.
La tecla de salida 1 se ilumina en naranja para indicar el estado de encendido
4 - Ejemplos básicos de configuración
Configurar más opciones de salida
188 TGF4000 Series Quick Start Guide
Configurar más opciones de salida
Requisito
En los ejemplos de configuración previos se ha mostrado cómo usar el menú de salida para
configurar el nivel de la salida (amplitud más desplazamiento, o nivel alto más nivel bajo) y
encenderla o apagarla. Este ejemplo demuestra la configuración de la fase de salida, la polaridad
de salida, la impedancia de la carga y el intervalo automático de la tensión.
Estado inicial
Antes de empezar, restaure los valores por defecto del instrumento según lo descrito en la sección
Estado inicial
Abrir el menú de salida
Pulse la tecla variable Output.
4 - Ejemplos básicos de configuración
Configurar más opciones de salida
189 TGF4000 Series Quick Start Guide
Cambiar la fase de salida
Se seleccionará por defecto la tecla variable Phase.
Introduzca una fase de - 4 5 °
El ángulo de fase configurado es el punto del periodo de la forma de onda que coincide con el
flanco de activación o sincronismo; es decir, el punto del periodo en el que comienza la forma de
onda. Consecuentemente, al configurarse una fase negativa, la forma de onda se adelanta,
mientras que con una fase negativa se retrasa, con respecto a la activación o el sincronismo; la
forma de onda de la caja de gráficas se actualiza para mostrar este cambio.
La tecla variable Align (alinear) se emplea para realinear la fase al realizar cambios en la frecuencia.
Cambiar la polaridad de salida
Pulse la tecla variable Type (tipo) para invertir la polaridad de la salida.
Tenga en cuenta que, pulsando sucesivamente la tecla Type, se alterna entre polaridad normal e
invertida.
4 - Ejemplos básicos de configuración
Configurar más opciones de salida
190 TGF4000 Series Quick Start Guide
Cambiar la impedancia de carga
Pulse la tecla variable Load (ca
rga).
Vuelva a pulsar la tecla variable Load para modificar la impedancia a High-z (alta
impedancia).
Advierta que, aunque el valor por defecto de la impedancia de carga es 50 ohmios, puede
cambiarse a cualquier otro entre 50 y 10.000 ohmios. Los niveles se calculan en función de esta
impedancia.
Al pulsar sucesivamente la tecla Load, su función alterna entre un valor numérico y High-z.
Observe cómo la lectura de la amplitud aumenta hasta 2 voltios pico a pico.
Pulse la tecla variable Load para devolver la impedancia de la carga a 50 ohmios.
4 - Ejemplos básicos de configuración
Configurar más opciones de salida
191 TGF4000 Series Quick Start Guide
Cambiar el intervalo
Pulse la tecla variable Range
(intervalo).
Vuelva a pulsar la tecla variable Range para modificar el intervalo de Auto (automático) a
Hold (fijo).
El modo Auto varía automáticamente el intervalo en pasos de 6 dB de atenuación (es decir,
máximos de intervalo de 10 Vpp, 5 Vpp, 2,5 Vpp, etc., en 50 con el intervalo de calibrado de la
amplitud limitado a 6 dB para mantener la calidad de la forma de onda.
Al seleccionar el modo Hold se deshabilita el intervalo automático; el valor de atenuación pasa a
ser fijo y el intervalo de calibrado de la amplitud deja de estar limitado.
Con el intervalo configurado en Auto, la amplitud y los atenuadores cambian automáticamente
para lograr un rendimiento óptimo. Con el intervalo configurado en Hold, se utiliza un valor de
atenuación fijo para todos los valores de amplitud.
5 - Repaso de las capacidades del generador
Configurar una señal de onda arbitraria
192 TGF4000 Series Quick Start Guide
5. REPASO DE LAS CAPACIDADES DEL
GENERADOR
En los siguientes ejemplos se indica únicamente la configuración de los parámetros, así como los
nombres de las teclas relacionadas. Se presentan las formas de onda de salida resultantes junto a
la forma de onda de sincronismo o activación cuando proceda. La amplitud de la salida y la
configuración del desplazamiento son tan solo ejemplos que no es necesario observar.
Configurar una señal de onda arbitraria
Comience con el instrumento de nuevo en la configuración por defecto.
MENÚ
Tecla variable
Forma de onda arbitraria
Waves
MENÚ
TECLA FÍSICA
Forma de onda arbitraria
Arb
Parámetro
Tecla variable
Configuración
Selección de onda
Waves
Sinc
Frecuencia
-
10kHz
Amplitud
Ampl
2V
Desviación
Offset
5mV
Parámetro
TECLA FÍSICA
Configuración
Estado de salida
Output 1
On
5 - Repaso de las capacidades del generador
Configurar una forma de onda senoidal con modulación AM
193 TGF4000 Series Quick Start Guide
Configurar una forma de onda senoidal con modulación AM
Comience con el instrumento de nuevo en la configuración por defecto.
Parámetro
Tecla variable
Configuración
Frecuencia
-
10MHz
MENÚ
TECLA FÍSICA
Modulación
Mod
Parámetro
Tecla variable
Configuración
Frecuencia
-
100kHz
Profundidad
Depth
100%
Fuente
Source
Internal
Forma
Shape
Sine
Estado de modulación
On/Off
On
Parámetro
TECLA FÍSICA
Configuración
Estado de salida
Output 1
On
Cambiar la modulación a AM-SC
Parámetro
Tecla variable
Configuración
Tipo
Type > AM
AM-SC
5 - Repaso de las capacidades del generador
SBSA
194 TGF4000 Series Quick Start Guide
SBSA
Comience con el instrumento de nuevo en la configuración por defecto.
MENÚ
TECLA FÍSICA
Forma de onda arbitraria
Waves
MENÚ
Tecla variable
Configuración
Ruido/SBSA
Noise
Fuente
Source
PRBS
Parámetro
Tecla variable
Configuración
Tasa de bits
BitRate
1Mbps
Amplitud
Ampl
3.3V
Desviación
Offset
1.65V
Tipo de SBSA
Type
PN7
Parámetro
TECLA FÍSICA
Configuración
Estado de salida
Output 1
On
5 - Repaso de las capacidades del generador
Modulación de frecuencia de una forma de onda de senoidal
195 TGF4000 Series Quick Start Guide
Modulación de frecuencia de una forma de onda de senoidal
Comience con el instrumento de nuevo en la configuración por defecto.
MENÚ
TECLA FÍSICA
Modulación
Mod
Parámetro
Tecla variable
Configuración
Estado de modulación
On/Off
On
Tipo de modulación
Type
FM
Frecuencia de modulación
-
1kHz
Desviación
Deviatn
9kHz
MENÚ
TECLA FÍSICA
Senoidal
Params
Parámetro
Tecla variable
Configuración
Amplitud
Ampl
1.0V
Desviación
Offset
0.0V
Parámetro
TECLA FÍSICA
Configuración
Estado de salida
Output 1
On
5 - Repaso de las capacidades del generador
Forma de onda modulada por anchura de pulso (PWM)
196 TGF4000 Series Quick Start Guide
Forma de onda modulada por anchura de pulso (PWM)
Comience con el instrumento de nuevo en la configuración por defecto.
MENÚ
TECLA FÍSICA
Forma de onda arbitraria
Waves
Parámetro
Tecla variable
Configuración
Pulso
Pulse
MENÚ
TECLA FÍSICA
Modulación
Mod
Parámetro
Tecla variable
Configuración
Estado de modulación
On/Off
On
Tipo de modulación
Type
PWM
Frecuencia de modulación
-
1kHz
Desviación
Dev %
40%
MENÚ
TECLA FÍSICA
Pulso
Params
Parámetro
Tecla variable
Configuración
Amplitud
Ampl
1.0V
Desviación
Offset
0.0V
Parámetro
TECLA FÍSICA
Configuración
Estado de salida
Output 1
On
5 - Repaso de las capacidades del generador
Modulación por desplazamiento de amplitud (ASK)
197 TGF4000 Series Quick Start Guide
Modulación por desplazamiento de amplitud (ASK)
Comience con el instrumento de nuevo en la configuración por defecto.
MENÚ
TECLA FÍSICA
Modulación
Mod
Parámetro
Tecla variable
Configuración
Estado de modulación
On/Off
On
Tipo de modulación
Type
ASK
Fuente de modulación
Source
Internal
Amplitud de salto
HpAmpl
100mV
Tasa de conmutación
Rate
1kHz
Polaridad del salto
HopPol
Positive
MENÚ
TECLA FÍSICA
Senoidal
Params
Parámetro
Tecla variable
Configuración
Amplitud
Ampl
1.0V
Desviación
Offset
0.0V
Parámetro
TECLA FÍSICA
Configuración
Estado de salida
Output 1
On
Advierta que la segunda traza representa la salida de la toma Main Output 2 cuando el canal 2 está
sincronizado: dicha salida sigue la señal de la onda moduladora.
5 - Repaso de las capacidades del generador
Barrido de frecuencia de onda senoidal
198 TGF4000 Series Quick Start Guide
Barrido de frecuencia de onda senoidal
Comience con el instrumento de nuevo en la configuración por defecto.
MENÚ
TECLA FÍSICA
Barrido
Sweep
Parámetro
Tecla variable
Configuración
Estado de barrido
On/Off
On
Frecuencia de paro
Freq > Stop
100kHz
MENÚ
TECLA FÍSICA
Senoidal
Params
Parámetro
Tecla variable
Configuración
Amplitud
Ampl
1.0V
Desviación
Offset
0.0V
Parámetro
TECLA FÍSICA
Configuración
Estado de salida
Output 1
On
5 - Repaso de las capacidades del generador
Generación de una ráfaga activada
199 TGF4000 Series Quick Start Guide
Generación de una ráfaga activada
Comience con el instrumento de nuevo en la configuración por defecto.
Parámetro
Tecla variable
Configuración
Frecuencia
-
6MHz
MENÚ
TECLA FÍSICA
Ráfaga
Burst
Parámetro
Tecla variable
Configuración
Estado de ráfaga
On/Off
On
Recuento de ráfaga
Count
3
Fuente de activación
SetTrg > Source > Int
Internal Trigger
Periodo de activación
SetTrg > Period
5ms
MENÚ
TECLA FÍSICA
Senoidal
Params
Parámetro
Tecla variable
Configuración
Amplitud
Ampl
1.0V
Desviación
Offset
0.0V
Parámetro
TECLA FÍSICA
Configuración
Estado de salida
Output 1
On
Advierta que la segunda traza representa la salida de la toma Main Output 2 cuando el canal 2 está
sincronizado: dicha salida sigue la señal de entrada de activación.
5 - Repaso de las capacidades del generador
Emparejamiento de la frecuencia de ambos canales
200 TGF4000 Series Quick Start Guide
Emparejamiento de la frecuencia de ambos canales
Comience con el instrumento de nuevo en la configuración por defecto.
MENÚ
TECLA FÍSICA
Utilidad
Utility
Parámetro
Tecla variable
Configuración
Frecuencias
Dual Ch >Freq
Coupled
Parámetro
TECLA FÍSICA
Configuración
Estado de salida
Output1
On
Estado de salida
Output2
On
MENÚ
TECLA FÍSICA
Senoidal
Params
Parámetro
Tecla variable
Configuración
Desplazamiento de fase
Output > Phase > Done
90°
Frecuencia
Freq
1MHz
Observe cómo, al configurar el canal 1 a 1MHz, el canal 2 también se configura a 1mHz.
Puede verse el desplazamiento de fase de 90 grados entre canales.
5 - Repaso de las capacidades del generador
Contador de frecuencia
201 TGF4000 Series Quick Start Guide
Contador de frecuencia
Comience con el instrumento de nuevo en la configuración por defecto.
MENÚ
TECLA FÍSICA
Utilidad
Utility
Parámetro
Tecla variable
Configuración
Contador
Instr > FrCntr > On/Off
Counter enabled
Fuente
Source
TRIG IN - DC coupled
Tipo
Freq
Frequency
Medición
Count
La caja de edición muestra la medición actual. Cuando no existe ninguna señal presente en la
entrada seleccionada, el contador indica No signal (sin señal) hasta que se aplique la señal
adecuada.
Cuando se aplica la señal adecuada, el contador mide y muestra de forma continua su lectura en la
caja de edición.
6 - Mantenimiento
Limpieza
202 TGF4000 Series Quick Start Guide
6. MANTENIMIENTO
Los fabricantes o sus agentes en el extranjero ofrecen un servicio de reparación para toda unidad
que desarrolle un defecto. Si los propietarios desearan establecer su propio servicio, esto sólo debe
realizarse por personas cualificadas en conjunto con el manual de servicio que puede adquirirse
directamente del Fabricante o de sus agentes en el extranjero.
Limpieza
Si la unidad de suministro de fuerza necesita ser limpiada, utilizar un paño brevemente
humedecido en agua o en un detergente suave. La ventana de visualización debe lustrarse con un
paño suave y seco.
7 - Specification
203 TGF4000 Series Quick Start Guide
7. SPECIFICATION
MODEL:
TGF4042 TGF4082 TGF4162 TGF4242
Waveforms
Standard waveforms:
Sine, Square, Ramp (Variable Symmetry), Triangle (50% Ramp symmetry), Positive Ramp (100% Ramp
symmetry), Negative Ramp (0% Ramp symmetry), Pulse, Noise (Gaussian), DC, Sin(x)/x, Exponential
Rise, Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Cardiac, Gaussian, Lorentz, D-
Lorentz and 4 User Defined Arbitrary Waveforms. Dozens of useful pre-built arbitrary waveforms are
also supplied on the website:
www.aimtti.com
-
PRBS
Sine
Frequency range:
1µHz to 40MHz
1µHz to 80MHz
1µHz to 160MHz
1µHz to 240MHz
Frequency resolution:
1
µ
Hz, 14 digits
1
µ
Hz, 15 digits
Output level
(into 50Ω):
10mVp-p to 10Vp
≤50MHz
10mVp-p to 10Vp-p
≤80MHz
10mVp-p to 5Vp-p
10mVp-p to 10Vp-p
10mVp-p to 10Vp-p
≤120MHz
10mVp-p to 5Vp-p
10mVp-p to 5Vp-p
≤240MHz
10mVp-p to 2.5Vp-p
10mVp-p to 2.5Vp-p
Amplitude flatness
(1Vp-p relative to
10 kHz):
≤10MHz
±0.1dB
≤100MHz
±0.2dB
≤160MHz
±0.6dB
≤240MHz
±1.0dB
Harmonic distortion
(1Vp-p)
≤10MHz
-65dBc
≤50MHz
-50dBc
≤80MHz
-40dBc
≤130MHz
-35dBc
≤240MHz
-28dBc
Total harmonic distortion
DC to 20kHz (typical):
0.05%
Non-harmonic spurii:
-65dBc
Phase noise
(10MHz, 1Vp-p, 10kHz offset):
-113dBc/Hz
Square
Frequency range:
1µHz to 25MHz
1µHz to 100MHz
Frequency resolution:
1
µ
Hz, 14 digits
1
µ
Hz, 15 digits
Output level (into 50Ω):
≤50MHz
10mVp-p to 10Vp-p
10mVp-p to 10Vp-p
≤100MHz
10mVp-p to 4Vp-p
Duty cycle:
0.001% to 99.999%, 0.001% resolution
Rise and fall times
(typical):
4p-p
10ns, fixed
3ns, fixed
4p-p
5ns, fixed
Aberrations (typical):
±5% of amplitude
Jitter (RMS):
<30ps (cycle to cycle)
Ramp
Frequency range:
1µHz to 5MHz
Frequency resolution:
1µHz, 13 digits
Output level (into 50Ω):
10mVp-p to 10Vp-p
Linearity error:
<0.1% to 100kHz
<0.1% to 200kHz
Variable symmetry:
0.00% to 100.00%, 0.01% resolution
Pulse
Frequency range:
1mHz to 25MHz
1mHz to 100MHz
Frequency Resolution:
1mHz, 11 digits
1mHz, 12 digits
Output level (into 50Ω):
50MHz
10mVp-p to 10Vpp
10mVp-p to 10Vpp
100MHz
10mVp-p to 4Vp-p
Aberrations (Typical):
±5% of amplitude (for transition time 10ns)
±3% of amplitude (for transition time 20ns)
<±2% of amplitude (for transition time > 40ns)
±5% of amplitude (for transition time 5ns)
±3% of amplitude (for transition time 10ns)
<±2% of amplitude (for transition time > 20ns)
7 - Specification
204 TGF4000 Series Quick Start Guide
MODEL:
TGF4042 TGF4082 TGF4162 TGF4242
Jitter RMS:
<30ps (cycle to cycle)
Rise and Fall Times:
Range: ≤ 4Vp-p
8ns to 799.999999984s (10% to 90%)
3ns to 799.999999989s (10% to 90%)
Range: > 4Vp-p
5ns to 799.999999989s (10% to 90%)
Resolution:
100ps
Accuracy:
±500ps ±0.01% of period
Rise and Fall times can be independently varied or can be varied together simultaneously.
Width:
Range: ≤ 4Vp-p
20ns to 999.999999980s
5ns to 999.999999995s
Range: > 4Vp-p
10ns to 999.999999990s
Resolution:
100ps
Accuracy:
±200ps ±0.01% of period
Duty:
0.001% to 99.999%, 0.01% resolution
Delay:
Range:
0ns to 999.99999996s
0ns to 999.99999998s
Resolution:
100ps
Accuracy:
±200ps ±0.01% of period
Delay can be entered as absolute delay or phase.
Arbitrary
In built arbitrary waveforms (Sin(x)/x, Exponential Rise, Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Cardiac, Gaussian, Lorentz and D-
Lorentz). Up to 4 user-defined waveforms may be stored in non-volatile memory. Waveforms can be defined by downloading of waveform data via
remote interfaces or from the instrument’s front panel.
Waveform Memory Size:
8192 points
Vertical Resolution:
14 bits
16 bits
Frequency Range:
In built
1µHz to 2MHz
1
µ
Hz to 4MHz
User defined
1µHz to 40MHz
1µHz to 80MHz
Frequency Resolution:
In built
1µHz, 13 digits
User defined
1µHz, 14 digits
Output Level (into 50Ω):
10mVp-p to 10Vp-p
Sampling rate:
400Msa/s
800MSa/s
Point to Point Jitter (Typical):
2.5ns
1.25ns
Rise and Fall Times: <8ns
<5ns for 100MHz filter
<8ns for 62.5MHz filter
Effective Analogue Bandwidth (-3dB):
50MHz
62.5MHz, 100MHz, User Selectable
Noise
Gaussian White Noise: Noise can also be used as modulating waveform.
Bandwidth (-3dB):
50MHz
100MHz
Noise crest factor (Vp/Vrms):
6.4
5.16
Output Level (into 50Ω):
10mVp-p to 10Vpp
PRBS (TGF4162 & TGF4242 only)
Bit Rate
-
1µbps to 100Mbps, 1µbps resolution
Sequence Length:
-
2m 1, where m = 7, 9, 11, 15, 20, 23, 29 or 31
Rise and Fall Times
(Typical):
≤ 4Vp-p
-
3ns, Fixed
> 4Vp-p
-
5ns, Fixed
Output Level:
-
10mVp-p to 10Vpp into 50Ω
Harmonic Output (TGF4162 & TGF4242 only)
Harmonic waveforms can be defined and stored in user-defined arbitrary waveform locations.
Frequency Range:
-
1µHz to 80MHz
Frequency Resolution:
-
1µHz, 14 digits
Harmonic Order: -
1 to 50, Up to 16 different harmonics order can
be defined
Harmonic Amplitude: -
0.0% to 100.0% of output amplitude, 0.1%
resolution
Harmonic Phase:
-
-360.0 to +360.0 degrees, 0.1 degree resolution
Output level (into 50Ω):
-
10mVp-p to 10Vp-p
Internal frequency reference
Internal Modulating Frequency:
<± 1ppm
Amplitude Depth:
<± 1ppm first year
Internal Modulating Frequency:
<1ppm over the specified temperature range
7 - Specification
205 TGF4000 Series Quick Start Guide
MODEL:
TGF4042 TGF4082 TGF4162 TGF4242
Modulation
AM (Amplitude Modulation) Normal & Suppressed Carrier
Carrier Waveforms:
Sine, Square, Ramp, Pulse, Noise, Arb
-
PRBS
Maximum Carrier Frequency:
25MHz, subject to carrier waveform
50MHz, subject to carrier waveform.
Modulation Source:
Internal/External
Internal Modulating Waveforms:
Sine, Square, Positive Ramp, Negative Ramp, Triangle, Gaussian Noise, DC, Sinc, Exponential Rise,
Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Gaussian, Lorentz, D-Lorentz, Cardiac
and User Defined Arbs
-
PRBS-PN7, PN9, PN11, PN15, PN20, PN23, PN29,
PN31
Internal Modulating Frequency:
1µHz to 10MHz, 1µHz resolution
Amplitude Depth:
0.00% to 100.00%, 0.01% resolution
FM (frequency modulation)
Carrier Waveforms:
Sine, Square, Ramp, Arb
Modulation Source:
Internal/External
Internal Modulating Waveforms:
Sine, Square, Positive Ramp, Negative Ramp, Triangle, Gaussian Noise, DC, Sinc, Exponential Rise,
Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Gaussian, Lorentz, D-Lorentz, Cardiac
and User Defined Arbs.
-
PRBS-PN7, PN9, PN11, PN15, PN20, PN23, PN29,
PN31
Internal Modulating Frequency:
1µHz to 10MHz, 1µHz resolution
Frequency Deviation:
DC to Fmax/2, 1µHz resolution
PM (phase modulation)
Carrier Waveforms:
Sine, Square, Ramp, Arb
Modulation Source:
Internal/External
Internal Modulating Waveforms:
Sine, Square, Positive Ramp, Negative Ramp, Triangle, Gaussian Noise, DC, Sinc, Exponential Rise,
Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Gaussian, Lorentz, D-Lorentz, Cardiac
and User Defined Arbs.
-
PRBS-PN7, PN9, PN11, PN15, PN20, PN23, PN29,
PN31
Internal Modulating Frequency:
1µHz to 10MHz, 1µHz resolution
Phase Deviation:
-360.000 to +360.000 degrees, 0.001 degree resolution
ASK (Amplitude Shift Keying)
Carrier Waveforms:
Sine, Square, Ramp, Pulse, Noise, Arb
-
PRBS
Maximum Carrier Frequency:
25MHz, subject to carrier waveform.
50MHz, subject to carrier waveform.
Source:
Internal/External (via TRIG IN)
Internal Modulation:
2mHz to 10MHz (50% duty cycle square)
FSK (Frequency Shift Keying)
Carrier Waveforms:
Sine, Square, Ramp, Arb
Source:
Internal/External (via TRIG IN)
Internal Modulation:
2mHz to 10MHz (50% duty cycle square)
BPSK (Binary Phase Shift Keying)
Carrier Waveforms:
Sine, Square, Ramp, Arb
Modulation Source:
Internal/External (via TRIG IN)
Internal Modulation:
2mHz to 10MHz (50% duty cycle square)
PWM (Pulse Width Modulation)
Carrier Waveforms:
Pulse
Modulation Source:
Internal/External
Internal Modulating Waveforms:
Sine, Square, Positive Ramp, Negative Ramp, Triangle, Gaussian Noise, DC, Sinc, Exponential Rise,
Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Gaussian, Lorentz, D-Lorentz, Cardiac,
and User Defined Arbs
-
PRBS-PN7, PN9, PN11, PN15, PN20, PN23, PN29,
PN31
Internal Modulating Frequency:
1
µ
Hz to 10MHz, 1
µ
Hz resolution
Pulse Width Deviation:
0% to 100% of pulse width, 0.01% resolution
7 - Specification
206 TGF4000 Series Quick Start Guide
MODEL:
TGF4042 TGF4082 TGF4162 TGF4242
SUM (Additive Modulation) (TGF4162 & TGF4242 only)
Carrier Waveforms:
-
Sine, Square, Ramp, Pulse, Noise, PRBS, Arb
Maximum Carrier Frequency:
-
50MHz, subject to carrier waveform.
Modulation Source:
-
Internal/External
Internal Modulating Waveforms: -
Sine, Square, Positive Ramp, Negative Ramp,
Triangle, Gaussian Noise, DC, Sinc, Exponential
Rise, Exponential Fall, Logarithmic Rise,
Logarithmic Fall, Haversine, Gaussian, Lorentz, D-
Lorentz, Cardiac, PRBS-PN7, PN9, PN11, PN15,
PN20, PN23, PN29, PN31 and User Defined Arbs.
Internal Modulating Frequency:
-
1
µ
Hz to 10MHz, 1
µ
Hz resolution
Ratio:
-
0% to 100%, 0.01% resolution
Triggered Burst
Each active edge of the trigger signal will produce one burst of the waveform.
Carrier Waveforms:
Sine, Square, Ramp, Pulse, Arb: A fixed number of cycles, specified as number of cycles are generated
at every trigger event.
Noise: Noise is reset to its start condition at every trigger event. Allows generating same random
noise sequence.
-
PRBS: A fixed number of bits, specified as
number of cycles are generated at every trigger
event
Maximum Carrier Frequency:
25MHz (finite cycles), Fmax(infinite), subject to
carrier waveform.
50MHz (finite cycles), Fmax(infinite), subject to
carrier waveform.
Number of Cycles:
1 to 2147483647 and infinite.
Trigger Repetition
Rate:
Internal
2mHz to 25MHz
2mHz to 50MHz
External
DC to 1MHz
Trigger Signal
Source:
Internal
from keyboard or trigger generator.
External
from TRIG IN or remote interface.
Trigger Start/Stop Phase:
-360.000 to +360.000 degrees, 0.001 degree resolution. Phase off-set cannot be set for Noise and
PRBS waveforms.
Gated
Waveform will run while the Gate signal is true and stop while false.
Carrier Waveforms:
Sine, Square, Ramp, Pulse, Noise, Arb
-
PRBS
Maximum Carrier Frequency:
25MHz, subject to carrier waveform.
50MHz, subject to carrier waveform.
Trigger Repetition
Rate:
Internal
2mHz to 25MHz
2mHz to 50MHz
External
DC to 1MHz
Gate Signal Source:
Internal
from keyboard or trigger generator.
External
from TRIG IN or remote interface.
Gate Start/Stop Phase:
-360.000 to +360.000 degrees, 0.001 degree resolution. Phase offset cannot be set for Noise and
PRBS waveforms
Sweep
Frequency sweep capability is provided for both standard and arbitrary waveforms
Carrier Waveforms:
Sine, Square, Ramp, Arb
Sweep Mode:
Linear or logarithmic, triggered or continuous.
Sweep Direction:
Up or Down
Sweep Range:
From 1
µ
Hz to Fmax. Phase continuous. Independent setting of the start and stop frequency.
Sweep Time:
1
µ
s to 500s (9 digit resolution).
Sweep Trigger Source:
The sweep may be free run or triggered from the following sources: Internal from keyboard or trigger
generator.
Externally from TRIG IN input or remote interface. External trigger repetition rate: DC to 1MHz
7 - Specification
207 TGF4000 Series Quick Start Guide
MODEL:
TGF4042 TGF4082 TGF4162 TGF4242
Trigger Generator
Internal Source Rate or Frequency:
20ns to 500s or 2mHz to 50MHz square waves
adjustable in 10ns steps.
40ns to 500s or 2mHz to 25MHz square waves
adjustable in 10ns steps.
Resolution:
10ns, 11 digits
Each channel has its own trigger generator. Channel 1 trigger is available for external use from the MAIN OUT 2 socket when Channel 2 is configured to
output Channel 1 sync waveform and sync source is set to trigger.
Dual-channel operations
Tracking
Independent (Off):
The channels are independent of each other.
Equal:
The two channels are identical and behave identically.
Coupling
Frequency coupling:
The frequencies of the two channels can be coupled. Changing the frequency of one channel changes
the frequencies of both channels.
Amplitude (and DC Offset) coupling:
Amplitude (and DC offset) of the two channels can be coupled. Changing the amplitude and offset on
one channel changes the amplitude and offset of both channels.
Output coupling:
Output On/Off can be coupled. Switching the output On/Off on one channel switches the output
On/Off of both channels.
Characteristics
Relative phase:
-360.000 to 360.000 degrees, 0.001 degree resolution (Phase offset cannot be set for Noise)
Channel to channel Skew (typical):
<1ns (when performing identical operations)
Crosstalk (typical):
<-80db
External Frequency Measurement
Function:
Frequency, Period, Positive Width, Negative Width, Duty Cycle
Frequency Range:
AC coupled
3Hz to >125MHz
DC coupled
100mHz to >125MHz
Source:
AC coupled
REF / COUNT (AC) IN
DC coupled
TRIG / COUNT (DC) IN
Frequency Resolution:
Up to 7 digits displayed.
Measurement Time:
Automatic
Input Range and
Sensitivity:
AC coupled
≤50MHz 100mVpp - 5Vpp
>50MHz 250mVpp - 5Vpp
Maximum input ±10V
DC coupled
Threshold typically 1.2V;
Sensitivity 100mVpp (≤50MHz), 250mVpp (<50MHz)
maximum input +5V, -1V
Hysteresis:
Input hysteresis voltage
10mV
Accuracy:
±1 digit ± time base accuracy.
Time base Accuracy:
<± 1ppm initial settling error, <± 1ppm oscillator ageing rate in the first year, <1ppm over the
specified temperature range
7 - Specification
208 TGF4000 Series Quick Start Guide
MODEL:
TGF4042 TGF4082 TGF4162 TGF4242
Outputs
Main Outputs
Output Impedance:
50
Amplitude
(Sine):
20mVp-p to 20Vp-p
open circuit, 10mVp-p
to 10Vp-p into 50
50MHz
20mVp-p to 20Vp-p
open circuit, 10mVp-p
to 10Vp-p into 50
80MHz
20mVp-p to 10Vp-p
open circuit, 10mVp-p
to 5Vp-p into 50
20mVp-p to 20Vp-p open circuit, 10mVp-p to
10Vp-p into 50
120MHz
20mVp-p to 10Vp-p open circuit, 10mVp-p to
5Vp-p into 50
240MHz
20mVp-p to 5Vp-p
open circuit, 10mVp-p
to 2.5Vp-p into 50
Amplitude
(Pulse):
20mVp-p to 20Vp-p
open circuit, 10mVp-p
to 10Vp-p into 50Ω
≤ 50MHz
20mVp-p to 20Vp-p open circuit, 10mVp-p to
10Vp-p into 50Ω
100MHz
20mVp-p to 8Vp-p open circuit, 10mVp-p to 4Vp-
p into 50Ω
Amplitude can be specified open circuit (hi Z) or into an assumed load of 1Ω to 10kΩ in Vpp
Amplitude Accuracy:
1.5% ±5mV at 1kHz into 50
DC Offset Range:
±10V. DC offset plus signal peak limited to ±10V from 50.
DC Offset Accuracy:
Typically 1% ±50mV.
Resolution:
3 digits or 1mV for both Amplitude and DC Offset.
Sync Output
Channel 2 can be configured to output Channel 1 sync from its MAIN OUT 2 socket. Sync is a multi-function output which is automatically selected to be
any of the following. Alternatively, the user can choose Sync to always be carrier referenced, to output the currently used trigger signal or turn it off.
Carrier Waveform
Sync:
Sine / Square / Ramp /
Pulse / Arbs
≤ 28.125MHz
A square wave with 50% duty cycle at the
waveform frequency.
≤ 80MHz
A sine wave at the waveform frequency.
62.5MHz
A square wave with 50% duty cycle at the
waveform frequency.
≤ 240MHz
A sine wave at the waveform frequency.
Pattern
A positive pulse which is 1 bit rate wide at the beginning of the sequence
Noise
No sync associated with noise.
Modulation Sync:
AM/FM/PM/SUM/
PWM
A square wave with 50% duty cycle referenced to the internal modulation waveform when
modulation source is internal, or a square wave referenced to the carrier waveform when
modulation source is external. No sync is associated with Noise and DC waveforms as the modulation
source.
ASK
A square wave referenced to the trigger rate. The sync is a TTL high when hop amplitude is the
output amplitude and TTL low when carrier amplitude is the output amplitude for positive slope and
vice versa for negative slope.
FSK
A square wave referenced to the trigger rate. The sync is a TTL high when hop frequency is the output
frequency and TTL low when carrier frequency is the output frequency for positive slope and vice
versa for negative slope.
BPSK
A square wave referenced to the trigger rate. The sync is a TTL high when the hop phase is the output
phase and TTL low when carrier phase is the output phase for positive slope and vice versa for
negative slope.
Sweep Sync:
A square wave that is a TTL high from the beginning of the sweep and a TTL low from the midpoint of
the sweep
Burst Sync:
Internal Trigger
A square wave with 50% duty cycle at the trigger frequency.
External Trigger
A square wave with same duty cycle and frequency as the external source.
Manual Trigger
A positive pulse which is approximately 18us wide at the beginning of the event.
Trigger:
Selects the current trigger signal
Output Signal Level:
Logic level nominally 3V
Output Impedance:
50
7 - Specification
209 TGF4000 Series Quick Start Guide
For details, request the EU Declaration of Conformity for this instrument via http://www.aimtti.com/support (serial no. needed).
General specifications apply for the temperature range 5°C to 40°C.
Accuracy specifications apply for the temperature range 18°C to 28°C after 30 minutes warm-up, at maximum output into 50Ω.
Typical specifications are determined by design and are not guaranteed.
MODEL:
TGF4042 TGF4082 TGF4162 TGF4242
Ref Clock Output
Buffered version of the 10MHz clock currently in use (internal or external)
Output Level:
Nominally 3V logic level from 50Ω.
Inputs
Trigger / Count (DC) Input
For ASK, FSK, BPSK, triggered sweep, gated burst, triggered burst and DC coupled external frequency measurement.
Frequency Range:
Trigger Input
DC 1MHz
Counter Input
100mHz to >125MHz
Signal Range:
Threshold typically 1.2V;
Sensitivity 100mVpp (≤50MHz), 250mVpp (>50MHz)
Maximum input +5V / -1V.
Minimum Pulse Width (Trigger Input):
50ns
Polarity (Trigger Input):
Selectable as high/rising edge or low/falling edge.
Input Impedance:
10k
External Modulation Input
For AM, FM, PM, SUM and PWM
Voltage Range:
± 2.5V full scale
Input Impedance:
5k typical
Bandwidth:
DC to 5MHz
Ref Clock / Count (AC) Input
Input for an external 10MHz reference clock and AC coupled external frequency measurement.
Voltage Range:
≤50MHz 100mVpp 5Vpp
>50MHz 250mVpp 5Vpp
Maximum Voltage:
+10V
Minimum Voltage:
-10V
Interfaces
Full digital remote control facilities are available through LAN, USB and optional GPIB interfaces.
LAN Interface
Ethernet 100/10base T hardware connection. 1.5 LXI Device Specification 2016
USB Interface
Standard USB 2.0 hardware connection. Implemented as virtual-COM port.
USB Flash Drive
For waveform and set-up storage/recall.
GPIB (optional)
Conforming with IEEE488.1 and IEEE488.2
General
Display:
4.3 inch (10.9 cm) transflective backlit TFT LCD, 480 x 272 pixels, 262144 colours, adjustable
brightness and contrast.
Data Entry:
Keyboard selection of mode, waveform etc.; value entry direct by numeric keys or by rotary control.
Stored Settings:
Up to 9 complete instrument setups may be stored and recalled from
internal memory.
Size:
Bench Top: 97mm height; 250mm width; 295mm depth
Rack mount: 86.5mm (2U) height; 213.5mm (½rack) width; 269mm depth
Weight:
3.2kg
Power:
100-240VAC ±10% 50/60Hz ; 100-120VAC ±10% 400Hz ; 60VA max. Installation Category II.
Operating Range:
+5°C to 40°C, 2080% RH.
Storage Range:
20°C to + 60°C.
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Options:
19 inch rack mounting kit.
Safety & EMC:
Complies with EN610101 & EN61326-1.
For details, request the EU Declaration of Conformity for this instrument via
http://www.aimtti.com/support (serial no. needed).
EXCELLENCE THROUGH EXPERIENCE
Aim-TTi is the trading name of Thurlby
Thandar Instruments Ltd. (TTi), one of
Europe’s leading manufacturers of test
and measurement instruments.
The company has wide experience in the
design and manufacture of advanced
test instruments and power supplies
built up over more than thirty years.
The company is based in the United
Kingdom, and all products are built at the
main facility in Hunngdon, close to the
famous university city of Cambridge.
TRACEABLE QUALITY SYSTEMS
TTi is an ISO9001 registered company operang
fully traceable quality systems for all processes
from design through to nal calibraon.
ISO9001:2015
Cercate number FM 20695
WHERE TO BUY AIM-TTI PRODUCTS
Aim-TTi products are widely available from
a network of distributors and agents in more
than sixty countries across the world.
To nd your local distributor, please visit our
website which provides full contact details.
Designed and built in Europe by:
Thurlby Thandar Instruments Ltd.
Glebe Road, Huntingdon, Cambridgeshire.
PE29 7DR United Kingdom
Tel: +44 (0)1480 412451 Fax: +44 (0)1480 450409
Email: [email protected] Web:
www.aimtti.com
*48591-10* *01*

Transcripción de documentos

QUICK START GUIDE EN FR DE IT ES TGF4000 SERIES 40MHz, 80MHz, 160MHz & 240MHz Dual Channel Arbitrary Function Generators CONTENTS Quick Guide in English ......................................................................................................................... 2 Guide Succinct en Français ................................................................................................................ 43 Kurzanleitung..................................................................................................................................... 83 Guida Rapida in Italiano ................................................................................................................... 123 Guía Rápida en Espaňol ................................................................................................................... 163 Specification .................................................................................................................................... 203 0 - Table of Contents TABLE OF CONTENTS 1. Introduction ................................................................................................................................................. 3 The TGF4000 Series of Arbitrary Function Generators ...................................................................................... 3 About this Guide ............................................................................................................................................... 3 2. Safety ........................................................................................................................................................... 4 General.............................................................................................................................................................. 4 Symbols ............................................................................................................................................................. 5 3. Operational Principles .................................................................................................................................. 6 Front Panel Layout ............................................................................................................................................ 6 Rear Panel Layout.............................................................................................................................................. 7 Screen Layout .................................................................................................................................................... 8 4. Getting Started............................................................................................................................................. 9 5. Basic Set-up Examples ................................................................................................................................ 10 Setting-up a Sine Wave Signal ......................................................................................................................... 10 Setting-up a Square Wave Clock Signal ........................................................................................................... 14 Setting-up a Pulse Waveform .......................................................................................................................... 20 Setting-up more Output Options .................................................................................................................... 28 Requirement ................................................................................................................................................... 28 6. Exploring the Generator Capabilities .......................................................................................................... 32 Setting-up an arbitrary wave signal ................................................................................................................. 32 Setting-up an AM modulated Sine Waveform ................................................................................................. 33 PRBS ................................................................................................................................................................ 34 Frequency Modulation of a Sine Waveform .................................................................................................... 35 Pulse Width Modulated Waveform (PWM) ..................................................................................................... 36 Amplitude shift keying (ASK) ........................................................................................................................... 37 Frequency Sweep of a Sine Wave.................................................................................................................... 38 Generating a Triggered Burst........................................................................................................................... 39 Coupling the Frequency of Both Channels ...................................................................................................... 40 Frequency counter .......................................................................................................................................... 41 7. Maintenance .............................................................................................................................................. 42 Cleaning .......................................................................................................................................................... 42 The latest revisions of this manual, device drivers and software tools can be downloaded from: http://www.aimtti.com/support 2 TGF4000 Series Quick Start Guide 1 - Introduction The TGF4000 Series of Arbitrary Function Generators 1. INTRODUCTION The TGF4000 Series of Arbitrary Function Generators This manual covers all four TGF4000 dual channel generators. Where there are differences in the specification, the limits for the TGF4042 & TGF4082 are shown in square brackets [ ] after the TGF4162 & TGF4242 limits. These programmable function/arbitrary generators use direct digital synthesis techniques to provide high performance and extensive facilities in a compact instrument. They generate a wide variety of waveforms with high resolution and accuracy. Sine waves are produced with low distortion to 160MHz/240MHz [40MHz/80MHz]. Square waves have fast rise and fall times at up to 100MHz [25MHz]. Linear ramp waves are produced to 5MHz. Ramp and square waves also have variable symmetry. The instruments generate high resolution, low jitter, variable edge time pulses to 100MHz [25MHz] with variable period, pulse width, pulse delay, pulse edges and amplitude. Complex custom waveforms can be generated with 16-bit [14-bit] resolution and a sampling rate of 800MSa/s [400MSa/s]. Up to four waveforms can be stored in internal memory. Waveforms can also be generated by the supplied Waveform Manager Plus V4.13 Windows application and downloaded to the instrument via USB, LAN or optional GPIB interfaces or via a USB flash drive. Front panel operation is straightforward and user friendly with all major parameters shown at all times on the large, bright, colour LCD. All major functions can be accessed with a single key or two. The knob or numeric keypad can be used to adjust frequency, amplitude, offset, and other parameters. Voltage values can be entered directly in Vpp or as high and low levels. Timing parameters can be entered in Hertz (Hz) or seconds. Internal AM, FM, PM, ASK, FSK, BPSK, SUM* and PWM modulation make it easy to modulate waveforms without the need for a separate modulation source. Linear and logarithmic sweeps are also built in, with sweep rates selectable from 1 µs to 500s. Burst mode operation allows for a userselected number of cycles at each trigger event. LAN and USB interfaces are standard and there is full compliance to 1.5 LXI Device Specification 2016. The instruments use a high stability temperature compensated internal oscillator and the external frequency reference input lets you synchronize to an external 10 MHz frequency standard for even greater accuracy. *TGF4162 & TGF4242 only About this Guide This Quick Start guide is for bench-top use of the TGF4000 Series comprising the TGF4042,TGF4082, TGF4162 and TGF4242 dual channel generators. A full Instruction Manual (English only) is also provided on the Aim-TTi website that includes comprehensive explanations of all functions and additional information on remote control, calibration, and the detailed technical specifications. 3 TGF4000 Series Quick Start Guide 2 - Safety General 2. SAFETY General This generator is a Safety Class I instrument according to IEC classification and has been designed to meet the requirements of EN61010−1 (Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use). It is an Installation Category II instrument intended for operation from a normal single phase supply. This instrument has been tested in accordance with EN61010−1 and has been supplied in a safe condition. This instruction manual contains some information and warnings which have to be followed by the user to ensure safe operation and to retain the instrument in a safe condition. This instrument has been designed for indoor use in a Pollution Degree 2 environment in the temperature range 5°C to 40°C, 20% − 80% RH (non−condensing). It may occasionally be subjected to temperatures between +5° and −10°C without degradation of its safety. Do not operate while condensation is present. Use of this instrument in a manner not specified by these instructions may impair the safety protection provided. Do not operate the instrument outside its rated supply voltages or environmental range. WARNING! THIS INSTRUMENT MUST BE EARTHED Any interruption of the mains earth conductor inside or outside the instrument will make the instrument dangerous. Intentional interruption is prohibited. The protective action must not be negated by the use of an extension cord without a protective conductor. When the instrument is connected to its supply, terminals may be live and opening the covers or removal of parts (except those to which access can be gained by hand) is likely to expose live parts. The apparatus shall be disconnected from all voltage sources before it is opened for any adjustment, replacement, maintenance or repair. Any adjustment, maintenance and repair of the opened instrument under voltage shall be avoided as far as possible and, if inevitable, shall be carried out only by a skilled person who is aware of the hazard involved. If the instrument is clearly defective, has been subject to mechanical damage, excessive moisture or chemical corrosion the safety protection may be impaired and the apparatus should be withdrawn from use and returned for checking and repair. Make sure that only fuses with the required rated current and of the specified type are used for replacement. The use of makeshift fuses and the short−circuiting of fuse holders is prohibited. This instrument uses a Lithium button cell for non−volatile memory battery back−up; typical life is 5 years. In the event of replacement becoming necessary, replace only with a cell of the correct type, i.e. 3V Li/Mn02 20mm button cell type 2032. Exhausted cells must be disposed of carefully in accordance with local regulations; do not cut open, incinerate, expose to temperatures above 60°C or attempt to recharge. Do not wet the instrument when cleaning it and in particular use only a soft dry cloth to clean the LCD window. 4 TGF4000 Series Quick Start Guide 2 - Safety Symbols Symbols The following symbols are used on the instrument and in this manual: Caution −refer to the accompanying documentation, incorrect operation may damage the instrument. Terminal connected to chassis ground. Mains supply OFF. l Mains supply ON. Alternating current. 5 TGF4000 Series Quick Start Guide 3 - Operational Principles Front Panel Layout 3. OPERATIONAL PRINCIPLES Front Panel Layout Ref. Short Description Function 1 Power Switch Switches instrument on or off. Safety Note: To fully disconnect from the AC supply, unplug the mains cord from the back of the instrument or switch off at the AC supply outlet; make sure that the means of disconnection is readily accessible. 2 Soft-keys Performs the function shown on the LCD soft-key label above. 3 Waveform Keys Selects the main waveform type (carrier waveform) as active. (Sine, Square, Ramp, Pulse, Noise/PBRS or Arb.) 4 Waveform Modification Menus Opens menus for setting parameters for Modulation, Sweep and Burst 5 Other Menus Selects menus for internal and external file storage, instrument utilities, and trigger conditions. 6 Main Sockets Main output sockets. Channel 2 can also be configured to output Channel 1 sync from its MAIN OUT 2 socket. 7 Output Keys Switch the selected MAIN OUT on or off. 8 Cursor Keys and Spin Wheel Used to change numeric parameter values digit by digit. Used to select items within some menus. 9 Numeric Keypad Used to enter numeric parameter values directly. 10 USB Flash Drive USB Host connector for USB Flash drive storage. 6 TGF4000 Series Quick Start Guide 3 - Operational Principles Rear Panel Layout Rear Panel Layout Ref. Short Description Function 1 Modulation Input Input for external modulation of main waveforms. 2 Reference In / AC coupled frequency counter Input for external 10MHz reference clock and AC coupled external frequency measurement. 3 Reference Out Output for internal 10MHz reference clock. 4 Trigger Input / DC coupled frequency counter Input for external triggering of main waveforms and DC coupled external frequency measurement 5 LAN connection Designed to meet LXI Core 2011.Remote control is possible using the TCP/IP Socket protocol. 6 USB connection Accepts a standard USB cable. 7 GPIB connection (optional) IEEE-488 The default GPIB address is 5. 7 TGF4000 Series Quick Start Guide 3 - Operational Principles Screen Layout Screen Layout Ref. Short Description Function 1 Channel Indicator Shows currently selected channel 2 Main Waveform type Shows current carrier waveform 3 Output State Shows main output On or Off 4 External Clock Indicator Shows status of external clock (if applied) 5 LAN Status Indicator Shows status of LAN (Ethernet) connection. 6 Parameters Box Shows main parameters for waveform. 7 Menu Description Shows the currently selected editing menu. 8 Graph Box Shows a graphical representation of the selected waveform. 9 Edit Box Shows the current parameter that can be edited 10 Soft-key Labels Shows the current functions for the six keys below. 8 TGF4000 Series Quick Start Guide 4 - Getting Started Initial Conditions 4. GETTING STARTED In order to familiarise the user with some of the basic functionalities of the instrument, a number of set-up examples are shown in this guide. It is recommended that all users should carry out the first four examples: • • • • Setting-up a Sine Wave Signal Setting-up a Square Wave Clock Signal Setting-up a Pulse Waveform Setting-up more Output Options A number of further set-up examples are provided that assume some familiarity with the instrument: • • Setting-up an arbitrary wave signal Setting-up an AM modulated Sine Waveform • • • • • • • • PRBS (TGF4162 & TGF4242) Frequency Modulation of a Sine Waveform Pulse Width Modulated Waveform (PWM) Amplitude shift keying (ASK) Frequency Sweep of a Sine Wave Generating a Triggered Burst Coupling the Frequency of Both Channels Frequency counter For more detailed information on all functionality - see the full Instruction Manual. Initial Conditions Before setting up the instrument for any of the examples, it should be returned to default conditions. To do this follow these steps: • • • • Press the hard key marked Utility Press the soft-key labelled System Press the soft-key labelled Default (display will show Restore Factory Default?) Press the soft-key labelled Yes This sets the main waveform to Sine (10kHz, 1V pk-pk) and cancels any modulations, sweep, or burst triggering or gating. NOTE The instrument can be set to remember its latest settings on power-off and restore them at poweron. This is set from the Utility > System menu and the PwrOn soft-key. This setting will be lost when the instrument is restored to default conditions as described above. 9 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Sine Wave Signal 5. BASIC SET-UP EXAMPLES Setting-up a Sine Wave Signal Requirement Output a continuous sine wave signal with 40MHz frequency and an amplitude of 6 volts pk-pk from MAIN OUT 1. Starting Conditions Before starting, reset the instrument to defaults as described in section 4 Getting Started Open Waveform Menu - Sine 10 • Press the hard key marked Waves • Press the soft key labelled Sine TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Sine Wave Signal Set the Frequency The soft key labelled Freq will be highlighted - the current frequency appears in the edit box. Pressing this soft-key repeatedly changes its function between Frequency and Period. • Use the numeric keypad to enter a new frequency. Press the numbers 4 0 As soon as a number is entered, the soft-keys change to show units of frequency. • 11 Press the soft-key labelled MHz to confirm a frequency of 40MHz. TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Sine Wave Signal Set the Amplitude • Press the soft key labelled Ampl Successive presses of the Ampl soft-key changes the Ampl and Offset key labels to HiLvl (high level) and LoLvl (low level) and vice versa. • Use the numeric keypad to enter a new amplitude. Press the number 6 As soon as a number is entered, the soft-keys change to show units of voltage. 12 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Sine Wave Signal • Press the soft-key labelled Vpp to confirm a pk-pk amplitude of 6.0 volts. Turn the Output On • Press Output 1 key to turn the channel 1 output On. The Output 1 key illuminates orange to indicate the on state. 13 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Square Wave Clock Signal Setting-up a Square Wave Clock Signal Requirement Output a continuous square wave clock signal with 20MHz frequency, 50% duty cycle and a high level of 3.3V and a low level of 0.0 volts from MAIN OUT 1. Starting Conditions Before starting, reset the instrument to defaults as described in section 4 Getting Started Open Waveform Menu - Square • Press the hard key marked Waves • Press the soft-key labelled Square. + 14 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Square Wave Clock Signal Set the Frequency The soft key labelled Freq will be highlighted - the current frequency appears in the edit box. Pressing this soft-key repeatedly changes its function between Frequency and Period. • Use the numeric keypad to enter a new frequency. Press the numbers 2 0 As soon as a number is entered, the soft-keys change to show units of frequency. • Press the soft-key labelled MHz to confirm a frequency of 20MHz. The graph box changes to show the rise time on the edges which is now significant. 15 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Square Wave Clock Signal Confirm the Duty Cycle • Press the soft-key labelled Duty - the current duty cycle appears in the edit box. The duty cycle is already set at 50%, but could be changed here if required. Set the High and Low Levels • Press the soft-key labelled Ampl - the key label changes to HiLvl and the current high level voltage appears in the edit box. Successive presses of the Ampl soft-key changes the Ampl and Offset key labels to HiLvl (high level) and LoLvl (low level) and vice versa. 16 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Square Wave Clock Signal When the soft-key is labelled HiLvl - the current high level voltage appears in the edit box. • Use the numeric keypad to enter a new level. Press the numbers 3 . 3 As soon as a number is entered, the soft-keys change to show units of voltage. • 17 Press the soft-key labelled V to confirm a high level of 3.3 volts. TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Square Wave Clock Signal • Press the soft-key labelled LoLvl - the current low level voltage appears in the edit box. • Use the numeric keypad to enter a new level. Press 0 • Press the soft-key labelled V to confirm a low level of 0.0 volts. Turn the Output On • Press Output 1 key to turn the channel 1 output On. The Output 1 key illuminates orange to indicate the on state 18 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Square Wave Clock Signal Making live changes to any numeric parameter (e.g. Frequency) Numeric parameters can be changed by using the cursor keys and spin wheel as an alternative to the numeric keypad. • Press the hard key marked Waves • Press the soft-key labelled Square. • Press the soft-key labelled Freq – the current frequency value of 20.0MHz is displayed • Press the Cursor hard keys to move the edit highlight to the second digit. • Use the spin wheel to change the value – the frequency is changed immediately. NOTE Press to activate/ deactivate the spin wheel. 19 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Pulse Waveform Setting-up a Pulse Waveform Requirement Output a continuous pulse signal with 100ns period, 30ns pulse width, 20ns edge times and a high level of 2.7V and a low level of -0.6 volts from MAIN OUT 1. Starting Conditions Before starting, reset the instrument to defaults as described in section 4 Getting Started Open Waveform Menu - Pulse 20 • Press the hard key marked Waves • Press the soft-key labelled Pulse. TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Pulse Waveform Set the Period • Press the soft-key labelled PlsFrq so that it changes to PlsPer - the current period appears in the edit box. Pressing this soft-key repeatedly changes its function between Frequency and Period. • Use the numeric keypad to enter a new period. Press the numbers 1 0 0 As soon as a number is entered, the soft-keys change to show units of time. 21 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Pulse Waveform • Press the soft-key labelled ns to confirm a period of 100ns. The graph box changes to show a representation of the pulse and edge times. Set the Pulse Width • 22 Press the soft-key labelled Duty - the key label changes to Width and displays the width as a time TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Pulse Waveform • Use the numeric keypad to enter a new width. Press the numbers 3 0. As soon as a number is entered, the soft-keys change to show units of time. • 23 Press the soft-key labelled ns to confirm a width of 30ns. TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Pulse Waveform Set the Pulse Edge Times • Press the soft-key labelled More • Press the soft-key labelled Edge • Use the cursor keys to select the digit representing units of 10ns • Use the spin wheel to change the value to 20.0ns The value could have been entered using the numeric keypad if preferred. • 24 Press the soft-key labelled Done TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Pulse Waveform Set the High and Low Levels • Press the soft-key labelled Ampl - the key label changes to HiLvl and the current high level voltage appears in the edit box Successive presses of the Ampl soft-key changes the Ampl and Offset key labels to HiLvl (high level) and LoLvl (low level) and vice versa. • Use the numeric keypad to enter a new level. Press the numbers 2 . 7 . As soon as a number is entered, the soft-keys change to show units of voltage. 25 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Pulse Waveform 26 • Press the soft-key labelled V to confirm a high level of 2.7 volts. • Press the soft-key labelled LoLvl - the current low level voltage appears in the edit box. TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up a Pulse Waveform • Use the numeric keypad to enter a new level. Press - . 6 • Press the soft-key labelled V to confirm a low level of -600 mV. Turn the Output On • Press Output1 key to turn the channel 1 output On. The Output 1 key illuminates orange to indicate the on state 27 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Setting-up more Output Options Setting-up more Output Options Requirement In the earlier set-up examples it was shown how the output menu is used to set the output level (amplitude plus offset or high level plus low level) and turn the output on or off. This example demonstrates the setting of output phase, output polarity, load impedance and voltage auto-range. Starting Conditions Before starting, reset the instrument to defaults as described in section 4 Getting Started Open the Output Menu • 28 Press the soft-key marked Output TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Requirement Change the Output Phase The soft-key labelled Phase will be selected as default. • Enter a phase of -45 degree. The set phase angle is the point in the waveform period which is coincident with the Sync or trigger edge, i.e. it is the point in the period at which the waveform starts. Hence a negative phase setting advances, and a positive phase setting delays the waveform relative to the Sync or trigger; the waveform in the graph box changes to show this. The Align soft-key is used to re-align phase when making frequency changes. Change the Output Polarity • Press the soft-key labelled Type to invert the output polarity. Successive presses of the type key alternates between normal and inverted. 29 TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Requirement Change the Load Impedance • Press the soft-key labelled Load • Press the soft-key labelled Load again to change the load impedance to High-z (high impedance). The default load impedance is 50 Ohms, but that this could be changed to any impedance between 50 and 10,000 Ohms. Levels are calculated based upon this impedance. Successive presses of the Load key alternates between a numeric value and High-z. Note that the amplitude readout increases to 2 volts pk-pk. • 30 Press the soft-key labelled Load to return the load impedance to 50 Ohms. TGF4000 Series Quick Start Guide 5 - Basic Set-up Examples Requirement Changing the Range • Press the soft-key labelled Range • Press the soft-key labelled Range again to change the range from Auto to Hold. Auto mode auto-ranges in 6dB attenuator steps (i.e. ‘range’ maximums of 10Vpp, 5Vpp, 2.5Vpp, etc., into 50Ω), with the amplitude range limited to 6dB to maintain waveform quality. Selecting Hold mode disables auto-ranging; the attenuator setting is fixed and the amplitude range is no longer limited. With range set to Auto the amplitude and attenuators will switch automatically and optimal performance will be realised. With range set to Hold a fixed attenuator setting is used for all amplitude settings. 31 TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities Setting-up an arbitrary wave signal 6. EXPLORING THE GENERATOR CAPABILITIES In the following examples only the parameter settings are described, together with the related key names. The resultant output waveforms are shown, along with the sync or trigger waveform where relevant. Output amplitude and offset settings are examples only and need not be followed. Setting-up an arbitrary wave signal Start with the instrument returned to Default Settings. MENU HARD KEY NAME Waveform Waves MENU Arbitrary waveform Soft-key Name Arb Parameter Wave selection Soft-key Name Waves Setting Sinc Frequency - 10kHz Amplitude Ampl 2V Offset Offset 5mVdc Parameter Output State HARD KEY NAME Output1 Setting On 32 TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities Setting-up an AM modulated Sine Waveform Setting-up an AM modulated Sine Waveform Start with the instrument returned to Default Settings. . Parameter Soft-key Name Frequency - Setting 10MHz MENU Modulation HARD KEY NAME Mod Parameter Frequency Soft-key Name - Setting 100kHz Depth Depth 100% Source Source Internal Shape Shape Sine Modulation state On/Off On Parameter Output State HARD KEY NAME Output1 Setting On Change the modulation to AM-SC Parameter Soft-key Name Type Type > AM 33 Setting AM-SC TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities PRBS PRBS Start with the instrument returned to Default Settings. MENU Waveform HARD KEY NAME Waves Menu Noise/PRBS Soft-key name Noise Setting Source Source PBRS Parameter Bit Rate Soft-key Name BitRate Setting 1Mbps Amplitude Ampl 3.3V Offset Offset 1.65V PRBS Type Type PN7 Parameter Output State HARD KEY NAME Output1 Setting On 34 TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities Frequency Modulation of a Sine Waveform Frequency Modulation of a Sine Waveform Start with the instrument returned to Default Settings. MENU Modulation HARD KEY NAME Mod Parameter Modulation State Soft-key Name On/Off Setting On Modulation Type Type FM Modulation Frequency - 1kHz Deviation Deviatn 9kHz MENU Sine HARD KEY NAME Params Parameter Amplitude Soft-key Name Ampl Setting 1.0V Offset Offset 0.0V Parameter Output State HARD KEY NAME Output1 Setting On 35 TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities Pulse Width Modulated Waveform (PWM) Pulse Width Modulated Waveform (PWM) Start with the instrument returned to Default Settings. MENU Waveform HARD KEY NAME Waves Parameter Pulse Soft-key Name Pulse MENU Modulation HARD KEY NAME Mod Parameter Modulation State Soft-key Name On/Off Setting On Modulation Type Type PWM Modulation Frequency - 1kHz Deviation Dev % 40% MENU Pulse HARD KEY NAME Params Parameter Amplitude Soft-key Name Ampl Setting 1.0V Offset Offset 0.0V Parameter Output State HARD KEY NAME Output1 Setting On 36 Setting TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities Amplitude shift keying (ASK) Amplitude shift keying (ASK) Start with the instrument returned to Default Settings. MENU Modulation HARD KEY NAME Mod Parameter Modulation State Soft-key Name On/Off Setting On Modulation type Type ASK Modulation Source Source Internal Hop Amplitude HpAmpl 100mV Switching Rate Rate 1kHz Hop Polarity HopPol Positive MENU Sine HARD KEY NAME Params Parameter Amplitude Soft-key Name Ampl Setting 1.0V Offset Offset 0.0V Parameter Output State HARD KEY NAME Output1 Setting On Note that the second trace is the output from the main Output 2 socket when Channel 2 is synchronised which follows the modulating waveform signal. 37 TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities Frequency Sweep of a Sine Wave Frequency Sweep of a Sine Wave Start with the instrument returned to Default Settings. MENU Sweep HARD KEY NAME Sweep Parameter Sweep State Soft-key Name On/Off Setting On Stop Frequency Freq > Stop 100kHz MENU Sine HARD KEY NAME Params Parameter Amplitude Soft-key Name Ampl Setting 1.0V Offset Offset 0.0V Parameter Output State HARD KEY NAME Output1 Setting On 38 TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities Generating a Triggered Burst Generating a Triggered Burst Start with the instrument returned to Default Settings. Parameter Frequency Soft-key Name - Setting 6MHz MENU HARD KEY NAME Burst Burst Parameter Burst State Soft-key Name On/Off Setting On Burst Count Count 3 Trigger Source SetTrg > Source > Int Internal Trigger Trigger Period SetTrg > Period 5ms MENU Sine HARD KEY NAME Params Parameter Amplitude Soft-key Name Ampl Setting 1.0V Offset Offset 0.0V Parameter Output State HARD KEY NAME Output1 Setting On Note that the second trace is the output from the Main Output 2 socket when Channel 2 is synchronised which follows the trigger input signal. 39 TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities Coupling the Frequency of Both Channels Coupling the Frequency of Both Channels Start with the instrument returned to Default Settings. MENU Utility HARD KEY NAME Utility Parameter Frequencies Soft-key Name Dual Ch > Freq Setting Coupled Parameter Output State HARD KEY NAME Output 1 Setting On Output State Output 2 On MENU Sine HARD KEY NAME Params Parameter Phase Shift Soft-key Name Output > Phase > Done Setting 90 degrees Frequency Freq 1MHz Note that, when channel 1 is set to 1MHz, channel 2 is also set to 1MHz. The 90 degree phase shift between the channels can be seen. 40 TGF4000 Series Quick Start Guide 6 - Exploring the Generator Capabilities Frequency counter Frequency counter Start with the instrument returned to Default Settings. MENU Utility HARD KEY NAME Utility Parameter Counter Soft-key Name Instr > FrCntr > On/Off Setting Counter enabled Source Source TRIG IN - DC coupled Type Freq Frequency Measurement Count The Edit Box shows the current measurement. With no input signal at the selected input, the counter displays ‘No signal’ until an adequate input signal is applied. When an adequate input signal is applied, the counter constantly measures and displays the reading in the Edit Box. 41 TGF4000 Series Quick Start Guide 7 - Maintenance Cleaning 7. MAINTENANCE The Manufacturers or their agents overseas will provide a repair service for any unit developing a fault. Where owners wish to undertake their own maintenance work, this should only be done by skilled personnel in conjunction with the service guide which may be obtained directly from the Manufacturers or their agents overseas. Cleaning If the instrument requires cleaning use a cloth that is only lightly dampened with water or a mild detergent. WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE INSTRUMENT, NEVER ALLOW WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE NEVER CLEAN WITH SOLVENTS. 42 TGF4000 Series Quick Start Guide 0 - Table des matières TABLE DES MATIÈRES 1. Sécurité ............................................................................................................................................................ 44 Symboles .............................................................................................................................................................. 45 2. Principes de fonctionnement ........................................................................................................................... 46 Disposition du panneau avant .............................................................................................................................. 46 Disposition du panneau arrière ............................................................................................................................ 47 Disposition de l'écran ........................................................................................................................................... 48 3. Première utilisation .......................................................................................................................................... 49 Conditions initiales ............................................................................................................................................... 49 4. Exemples de configuration de base .................................................................................................................. 50 Configurer un signal d'onde sinusoïdale............................................................................................................... 50 Configurer un signal d'horloge à onde carrée ...................................................................................................... 54 Configuration d'une forme d'onde à impulsion .................................................................................................... 60 Configurer de plus amples options pour les sorties ............................................................................................. 68 5. Explorer les fonctionnalités du générateur ....................................................................................................... 72 Configurer un signal d'onde arbitraire.................................................................................................................. 72 Configurer une forme d'onde sinusoïdale modulée AM ...................................................................................... 73 PRBS ..................................................................................................................................................................... 74 Modulation de fréquence d'une forme d'onde sinusoïdale ................................................................................. 74 Forme d'onde modulée à largeur d'impulsion (PWM) ......................................................................................... 76 Modulation par déplacement d’amplitude (FSK) ................................................................................................. 77 Balayage de fréquence d'une onde sinusoïdale ................................................................................................... 78 Génération d'une rafale déclenchée .................................................................................................................... 78 Couplage des fréquences de deux voies .............................................................................................................. 80 Compteur de fréquence ....................................................................................................................................... 81 6. Maintenance .................................................................................................................................................... 82 Nettoyage ............................................................................................................................................................. 82 Les dernières versions de ce manuel, les pilotes de périphérique et les outils logiciels peuvent être téléchargés à partir de: www.aimtti.com 43 TGF4000 Series Quick Start Guide 1 - Sécurité 1. SÉCURITÉ Cet instrument est de Classe de sécurité 1 suivant la classification IEC et il a été construit pour satisfaire aux impératifs EN61010-1 (Impératifs de sécurité pour le matériel électrique en vue de mesure, commande et utilisation en laboratoire). Il s’agit d’un instrument d’installation Catégorie II devant être exploité depuis une alimentation monophasée habituelle. Cet instrument a été soumis à des essais conformément à EN61010-1 et il a été fourni en tout état de sécurité. Ce manuel d’instructions contient des informations et avertissements qui doivent être suivis par l’utilisateur afin d’assurer un fonctionnement en toute sécurité et de conserver l’instrument dans un état de bonne sécurité. Cet instrument a été conçu pour être utilisé en interne dans un environnement de pollution Degré 2, plage de températures 5°C à 40°C, 20% - 80% HR (sans condensation). Il peut être soumis de temps à autre à des températures comprises entre +5°C et –10°C sans dégradation de sa sécurité. Ne pas l’utiliser lorsqu’il y a de la condensation. Toute utilisation de cet instrument de manière non spécifiée par ces instructions risque d’affecter la protection de sécurité conférée. Ne pas utiliser l’instrument à l’extérieur des tensions d’alimentation nominales ou de la gamme des conditions ambiantes spécifiées. AVERTISSEMENT! CET INSTRUMENT DOIT ETRE RELIE A LA TERRE Toute interruption du conducteur de terre secteur à l’intérieur ou à l’extérieur de l’instrument rendra l’instrument dangereux. Il est absolument interdit d’effectuer une interruption à dessein. Ne pas utiliser de cordon de prolongation sans conducteur de protection, car ceci annulerait sa capacité de protection. Lorsque l’instrument est relié à son alimentation, il est possible que les bornes soient sous tension et par suite, l’ouverture des couvercles ou la dépose de pièces (à l’exception de celles auxquelles on peut accéder manuellement) risque de mettre à découvert des pièces sous tension. Il faut débrancher toute source de tension éventuelle de l’appareil avant de l’ouvrir pour effectuer des réglages, remplacements, travaux d’entretien ou de réparation. Eviter dans la mesure du possible d’effectuer des réglages, travaux de réparation ou d’entretien lorsque l’instrument ouvert est branché à une source d’alimentation, mais si c’est absolument nécessaire, seul un technicien compétent au courant des risques encourus doit effectuer ce genre de travaux. S’il est évident que l’instrument est défectueux, qu’il a été soumis à des dégâts mécaniques, à une humidité excessive ou à une corrosion chimique, la protection de sécurité sera amoindrie et il faut retirer l’appareil, afin qu’il ne soit pas utilisé, et le renvoyer en vue de vérifications et de réparations. Uniquement remplacer les fusibles par des fusibles d’intensité nominale requise et de type spécifié. Il est interdit d’utiliser des fusibles bricolés et de court-circuiter des porte-fusibles. L’instrument utilise une pile bouton au lithium pour la mémoire non-volatile ; sa durée de vie est environ 5 ans. Pour son remplacement, utilisé une pile du même type : 3V Li/ Mn02 type 2032. Les piles usées doivent être jetées en accord avec les lois locales ; ne pas les couper, les brûler, les exposer à des températures au delà de 60°C ou essayer de la recharger. Ne pas mouiller l’instrument lors de son nettoyage; en particulier, n’utiliser qu’un chiffon doux et sec pour nettoyer la vitre de l’afficheur. 44 TGF4000 Series Quick Start Guide 1 - Sécurité Symboles Symboles Les symboles suivants se trouvent sur l’instrument, ainsi que dans ce manuel. ATTENTION - se référer à la documentation ci-jointe; toute utilisation incorrecte risque d’endommager l’appareil. Borne reliée à la terre du châssis Alimentation secteur ON (allumée) l Alimentation secteur OFF (éteinte) Courant alternatif (c.a.) 45 TGF4000 Series Quick Start Guide 2 - Principes de fonctionnement Disposition du panneau avant 2. PRINCIPES DE FONCTIONNEMENT Disposition du panneau avant Réf. Brève description Fonction 1 Commutateur d'alimentation Allume et éteint l'appareil. Remarque de sécurité : Afin de déconnecter complètement l’appareil de l’alimentation CA, débrancher le cordon d’alimentation à l’arrière de l’appareil ou éteindre la prise d’alimentation CA ; il convient de s’assurer que l'élément utilisé pour déconnecter l'appareil est accessible. 2 Touches programmables Exécutent la fonction indiquée sur l'étiquette située au-dessus de la touche programmable sur l'écran LCD. 3 Sélection des formes d'ondes Ouvre les menus de formes d'ondes pour régler les paramètres des formes d'ondes.(sinusoïdales, carrées, rampes, impulsions, Bruit/PBRS ou arb.) 4 Menus de modification des formes d'onde Ouvre les menus pour définir les paramètres de modulation, de balayage et de rafales 5 Autres menus Sélectionne les menus relatifs aux espaces de stockage internes et externes, aux utilitaires de l'appareil et aux conditions de déclenchement. 6 Prises principales Prises de sortie principales La voie 2 peut également être configurée pour produire la synchronisation de la voie 1 à partir de sa prise MAIN OUT 2 (SORTIE PRINCIPALE 2). 7 Touches de sortie Active ou désactive la sortie principale MAIN OUT sélectionnée. 8 Touches du curseur et molette Utilisées pour changer les valeurs des paramètres numériques chiffre par chiffre. Utilisée pour sélectionner des éléments dans certains menus. 9 Pavé numérique Utilisé pour saisir les valeurs de paramètres numériques directement. 10 Clé USB Connecteur hôte USB pour disque de stockage USB. 46 TGF4000 Series Quick Start Guide 2 - Principes de fonctionnement Disposition du panneau arrière Disposition du panneau arrière Réf. Brève description Fonction 1 Entrée de modulation Entrée externe pour la modulation des formes d'ondes principales. 2 Entrée de référence / Compteur de fréquence de couplage CA Entrée pour l’horloge de référence externe de 10 MHz et la mesure de la fréquence externe de couplage CA. 3 Sortie de référence Entrée pour une horloge de référence interne de 10 MHz. 4 Entrée de déclenchement / Compteur de fréquence de couplage CC Entrée externe de déclenchement des formes d’ondes principales et de mesure de la fréquence externe de couplage CC 5 LAN L’interface LAN (réseau local) est conçue pour répondre aux critères LXI Core 2011. Il est possible d’utiliser l'interface LAN à distance grâce au protocole de prise TCP/IP. 6 USB Accepte un câble USB standard. 7 GPIB- En option L’adresse GPIB par défaut est 5. 47 TGF4000 Series Quick Start Guide 2 - Principes de fonctionnement Disposition de l'écran Disposition de l'écran Réf. Brève description Fonction 1 Indicateur de voie Affiche la voie actuellement sélectionnée 2 Type de forme d'onde Affiche la forme d'onde porteuse actuelle 3 État de sortie Indique si la sortie principale est activée (On) ou désactivée (Off) 4 Témoin d'horloge externe Affiche l'état de l'horloge externe (si demandé) 5 Indicateur d'état LAN Affiche l'état de connexion au réseau local (Ethernet). 6 Fenêtre Paramètres Affiche les principaux paramètres pour la forme d'onde. 7 Description du menu Affiche le menu d'édition actuellement sélectionné. 8 Fenêtre Graphique Affiche une représentation graphique de la forme d'onde sélectionnée. 9 Fenêtre Modification Affiche le paramètre actuel qui peut être modifié 10 Étiquette de touche programmable Affiche les fonctions actuelles pour les six touches situées en dessous. 48 TGF4000 Series Quick Start Guide 3 - Première utilisation Conditions initiales 3. PREMIÈRE UTILISATION Afin de familiariser l'utilisateur avec certaines des fonctionnalités fondamentales de l'instrument, ce guide comporte un certain nombre d'exemples de configuration. Il est recommandé à tous les utilisateurs d'exécuter les trois premiers exemples : • • • • Configurer un signal d'onde sinusoïdale Configurer un signal d'horloge à onde carrée Configuration d'une forme d'onde à impulsión Configurer de plus amples options pour les sorties Plusieurs autres exemples de configuration sont fournis et présument que l'utilisateur est familier avec l'appareil : • • • • • • • • • • Configurer un signal d'onde arbitraire Configurer une forme d'onde sinusoïdale modulée AM PRBS (TGF4162 & TGF4242) Modulation de fréquence d'une forme d'onde sinusoïdale Forme d'onde modulée à largeur d'impulsion (PWM) Modulation par déplacement d’amplitude (ASK) Balayage de fréquence d'une onde sinusoïdale Génération d'une rafale déclenchée Couplage des fréquences de deux voies Compteur de fréquence Conditions initiales Avant de configurer l'appareil pour l'un des exemples, il doit être retourné aux conditions par défaut. Pour ce faire, procédez comme suit : • • • • Appuyez sur la touche marquée Utility Appuyez sur la touche programmable étiquetée System Appuyez sur la touche programmable étiquetée Default (l'écran affichera Restore Factory Default?) Appuyez sur la touche programmable étiquetée Yes Ceci règle la principale forme d'onde sinusoïdale (10 kHz, 1 V de crête à crête) et annule tout déclenchement de modulations, balayages ou rafales. REMARQUE : L'appareil peut être réglé de façon à se souvenir de ses derniers paramètres au moment de la mise à l'arrêt et à les restaurer lors de la mise en marche. Ce réglage peut être configuré à partir du menu Utility avec la touche programmable PwrOn. Ce réglage sera perdu lorsque l'instrument sera réinitialisé aux conditions par défaut, comme décrit ci-dessus. 49 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'onde sinusoïdale 4. EXEMPLES DE CONFIGURATION DE BASE Configurer un signal d'onde sinusoïdale Exigence Émission d'un signal sinusoïdal continu d'une fréquence de 40 MHz et d'une amplitude de 6 V de crête à crête à partir de la sortie principale 1 (MAIN OUT 1). Conditions de démarrage Avant de commencer, réinitialiser l'appareil aux paramètres par défaut comme décrit dans la rubrique 3 Première utilisation 50 • Appuyez sur la touche marquée Waves • Appuyez sur la touche programmable étiquetée Sine TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'onde sinusoïdale Régler la fréquence La touche programmable étiquetée Freq est mise en surbrillance : la fréquence actuelle s'affiche dans la zone d'édition. Notez qu'en appuyant plusieurs fois sur cette touche programmable, sa fonction bascule entre Fréquence et Période. • Utiliser le pavé numérique pour saisir une nouvelle fréquence, Appuyer sur les chiffres 4 0 Veuillez noter que, dès qu'un nombre est saisi, les touches programmables changent pour afficher les unités de fréquence. • 51 Appuyer sur la touche programmable étiquetée MHz pour confirmer une fréquence 40 MHz. TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'onde sinusoïdale Régler l'amplitude • Appuyez sur la touche programmable étiquetée Ampl Veuillez noter que des pressions successives sur la touche programmable Ampl changent les étiquettes des touches Ampl et Offset qui deviennent HiLvl (niveau haut) et LoLvl (niveau bas) et vice versa. • Utiliser le pavé numérique pour saisir une nouvelle amplitude. Appuyer sur le chiffre 6 Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher des unités de tension. 52 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'onde sinusoïdale • Appuyer sur la touche programmable étiquetée Vpp pour confirmer une amplitude de crête à crête de 6,0 V. Activer la sortie • Appuyer sur la touche Output 1 (Sortie 1) pour activer la voie 1. La touche de sortie 1 s'allume en orange pour indiquer l'état activé 53 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'horloge à onde carrée Configurer un signal d'horloge à onde carrée Exigence Émission d'un signal d'horloge d'onde carrée continu d'une fréquence de 20 MHz, avec un cycle de service de 50 %, un niveau élevé de 3,3 V et un niveau bas de 0,0 V provenant de la sortie principale 1 (MAIN OUT 1). Conditions de démarrage Avant de commencer, réinitialiser l'appareil aux paramètres par défaut comme décrit dans la rubrique 3 Première utilisation Ouvrir le menu de forme d'onde - Carré 54 • Appuyer sur la touche marquée Waves • Appuyer sur la touche programmable étiquetée Square. TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'horloge à onde carrée Régler la fréquence La touche programmable étiquetée Freq est mise en surbrillance : la fréquence actuelle s'affiche dans la zone d'édition. Notez qu'en appuyant plusieurs fois sur cette touche programmable, sa fonction bascule entre Fréquence et Période. • Utiliser le pavé numérique pour saisir une nouvelle fréquence. Appuyer sur les chiffres 2 0 Veuillez noter que, dès qu'un nombre est saisi, les touches programmables changent pour afficher les unités de fréquence. • Appuyer sur la touche programmable étiquetée MHz pour confirmer une fréquence de 20 MHz. Notez que la fenêtre graphique change pour afficher le temps de montée sur les fronts, qui est maintenant significatif. 55 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'horloge à onde carrée Confirmer le cycle de service • Appuyer sur la touche programmable étiquetée Duty - le cycle de service actuel apparaît dans la fenêtre d'édition. Veuillez noter que le cycle de service est déjà réglé sur 50 %, mais qu'il peut être modifié ici le cas échéant. Régler les niveaux haut et bas • Appuyer sur la touche programmable étiquetée Ampl - l'étiquette de la touche devient HiLvl et la haute de niveau haut actuelle s'affiche dans la zone d'édition Veuillez noter que des pressions successives sur la touche programmable Ampl changent les étiquettes des touches Ampl et Offset qui deviennent HiLvl (niveau haut) et LoLvl (niveau bas) et vice versa. 56 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'horloge à onde carrée Lorsque la touche programmable est étiquetée HiLvl - la tension de niveau haut actuelle apparaît dans la fenêtre d'édition. • Utiliser le pavé numérique pour saisir un nouveau niveau. Appuyer sur les chiffres 3 . 3 Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher des unités de tension. • 57 Appuyer sur la touche programmable étiquetée V pour confirmer un niveau haut de 3.3 V. TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'horloge à onde carrée • Appuyer sur la touche programmable étiquetée LoLvl - la tension de niveau bas actuelle apparaît dans la fenêtre d'édition. • Utiliser le pavé numérique pour saisir un nouveau niveau. Appuyer sur • Appuyer sur la touche programmable étiquetée V pour confirmer un niveau bas de 0,0 V. 0 Activer la sortie • Appuyer sur la touche Output 1 (Sortie 1) pour activer la voie 1. La touche de sortie 1 s'allume en orange pour indiquer l'état activé 58 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer un signal d'horloge à onde carrée Apporter des modifications en direct à tout paramètre numérique (par ex. la fréquence) Les paramètres numériques peuvent être modifiés à l'aide des les touches du curseur et de la molette au lieu d'utiliser le pavé numérique. • Appuyer sur la touche marquée Waves • Appuyer sur la touche programmable étiquetée Square. • Appuyer sur la touche programmable étiquetée Freq - la valeur de fréquence actuelle de 20 MHz s'affiche • Appuyer sur les touches du curseur pour déplacer la surbrillance d'édition jusqu'au deuxième chiffre. • Utiliser la molette pour modifier la valeur - la fréquence change immédiatement. REMARQUE : Appuyez sur pour activer / désactiver la molette. 59 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configuration d'une forme d'onde à impulsión Configuration d'une forme d'onde à impulsión Exigence Émission d'un signal d'impulsion continu d'une période de 100 ns, avec une largeur d'impulsion de 30 ns, un temps de front de 20 ns, un niveau haut de 2,7 V et un niveau bas de -0,6 V provenant de la sortie principale 1 (MAIN OUT 1). Conditions de démarrage Avant de commencer, réinitialiser l'appareil aux paramètres par défaut comme décrit dans la rubrique 3 Première utilisation Ouvrir le menu de forme d'onde - Impulsion 60 • Appuyer sur la touche marquée Waves • Appuyer sur la touche programmable étiquetée Pulse. TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configuration d'une forme d'onde à impulsión Régler la période • Appuyer sur la touche programmable étiquetée PlsFrq de façon à ce qu'elle affiche PlsPer - la période actuelle s'affiche dans la zone d'édition. Notez qu'en appuyant plusieurs fois sur cette touche programmable, sa fonction bascule entre Fréquence et Période. • Utiliser le pavé numérique pour saisir une nouvelle période. Appuyer sur les chiffres 1 0 0 Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher des unités de temps. 61 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configuration d'une forme d'onde à impulsión • Appuyer sur la touche programmable étiquetée ns pour confirmer une période de 100 ns. Notez que la fenêtre graphique change pour afficher une représentation de l'impulsion et des temps de front. Régler la largeur d'impulsion • 62 Appuye sur la touche programmable étiquetée Duty - l’étiquette de la touche devient Width et affiche la largeur sous forme de durée TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configuration d'une forme d'onde à impulsión • Utiliser le pavé numérique pour saisir une nouvelle largeur. Appuyer sur les chiffres 3 0. Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher des unités de temps. • 63 Appuyer sur la touche programmable étiquetée ns pour confirmer une largeur de 30 ns. TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configuration d'une forme d'onde à impulsión Régler les temps de front d'impulsion • Appuyer sur la touche programmable étiquetée More • Appuyer sur la touche programmable étiquetée Edge • Utiliser les touches de curseur pour sélectionner le chiffre représentant des unités de 10.0 ns • Utiliser la molette pour modifier la valeur et la régler sur 20,0 ns Veuillez noter que la valeur aurait pu être saisie à l'aide du pavé numérique selon les préférences de chacun. • 64 Appuyer sur la touche programmable étiquetée Done TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configuration d'une forme d'onde à impulsión Régler les niveaux haut et bas • Appuyer sur la touche programmable étiquetée Ampl - l'étiquette de la touche devient HiLvl et la haute de niveau haut actuelle s'affiche dans la zone d'édition Veuillez noter que des pressions successives sur la touche programmable Ampl changent les étiquettes des touches Ampl et Offset qui deviennent HiLvl (niveau haut) et LoLvl (niveau bas) et vice versa. • Utiliser le pavé numérique pour saisir un nouveau niveau. Appuyer sur les chiffres 2 . 7 Veuillez noter que, dès qu'un chiffre est saisi, les touches programmables changent pour afficher des unités de tension. 65 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configuration d'une forme d'onde à impulsión 66 • Appuyer sur la touche programmable étiquetée V pour confirmer une tension haute de 2,7 V. • Appuyer sur la touche programmable étiquetée LoLvl - la tension de niveau bas actuelle apparaît dans la fenêtre d'édition. TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configuration d'une forme d'onde à impulsión • Utiliser le pavé numérique pour saisir un nouveau niveau. Appuyer sur - . 6 • Appuyer sur la touche programmable étiquetée V pour confirmer un niveau bas de-600 mV. Activer la sortie Appuyer sur la touche Output1 (Sortie 1) pour activer la voie 1. La touche de sortie 1 s'allume en orange pour indiquer l'état activé 67 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer de plus amples options pour les sorties Configurer de plus amples options pour les sorties Exigence Les exemples de configuration précédents ont montré comment utiliser le menu sortie pour définir le niveau de sortie (amplitude plus décalage ou niveau haut au plus niveau bas) et pour activer ou désactiver la sortie. Cet exemple présente le réglage de la phase de sortie, de la polarité de sortie, de l'impédance de charge et de la plage automatique de tension. Conditions de démarrage Avant de commencer, réinitialiser l'appareil aux paramètres par défaut comme décrit dans la rubrique 3 Première utilisation Ouvrir le Menu Output (Sortie) • 68 Appuyer sur la touche programmable marquée Output TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer de plus amples options pour les sorties Modifier la phase de sortie La touche programmable étiquetée Phase est sélectionnée par défaut. • Saisir une phase de -45 degrés. L'angle de phase réglé est le point de la période de la forme d'onde qui coïncide avec le front de synchronisation ou de déclenchement. Autrement dit, il s'agit du point de la période auquel commence la forme d'onde. C'est pourquoi un réglage de phase négatif avance, et un réglage de phase positif retarde la forme d'onde par rapport à la synchronisation ou au déclenchement ; la forme d'onde présentée dans la fenêtre graphique change pour illustrer cet état de fait. La touche programmable Align est utilisée pour réaligner la phase lorsque des modifications sont apportées à la fréquence. Modifier la polarité de sortie • Appuyer sur la touche programmable étiquetée Type pour inverser la polarité de sortie. Veuillez noter que des pressions successives de la touche Type permettent d'alterner entre une polarité normale et inversée. 69 TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer de plus amples options pour les sorties Modifier l'impédance de charge • Appuyez sur la touche programmable étiquetée Load • Appuyer à nouveau sur la touche programmable étiquetée Load pour modifier l'impédance de charge et la régler sur High-z. (Impédance élevée). Veuillez noter que, par défaut, l'impédance de charge est de 50 Ohms, mais que cette valeur peut être modifiée au profit de n'importe quelle impédance d'une valeur comprise entre 50 et 10 000 Ohms. Les niveaux sont calculés en fonction de cette impédance. Des pressions successives de la touche Load permettent d'alterner entre une valeur numérique et High-z. Veuillez noter que la mesure de l'amplitude augmente pour atteindre 2 V de crête à crête. • 70 Appuyer sur la touche programmable étiquetée Load pour ramener l'impédance de charge à 50 Ohms. TGF4000 Series Quick Start Guide 4 - Exemples de configuration de base Configurer de plus amples options pour les sorties Modifier la plage • Appuyez sur la touche programmable étiquetée range • Appuyer à nouveau sur la touche programmable étiquetée range pour faire passer la plage du mode Auto (automatique) au mode Hold (maintien). Le mode Auto règle la gamme automatiquement par pas de 6 dB d’atténuateur (c.-à-d. ‘gamme’ maximum de 10 Vpp, 5 Vpp, 2,5 Vpp, etc., à 50 Ω), avec une gamme du vernier de réglage de l’amplitude limitée à 6 dB afin de maintenir la qualité de la forme d’onde. La sélection du mode Hold désactive le réglage automatique de la gamme ; le réglage actuel de l’atténuateur est fixé et la gamme du vernier de réglage de l’amplitude n’est plus limitée. Lorsque la gamme est réglée sur Auto, l'amplitude et les atténuateurs sont automatiquement activés pour obtenir une performance optimale. Lorsque la plage est réglée sur Hold (Maintien), un réglage fixe de l’atténuateur est utilisé pour tous les réglages de l’amplitude. 71 TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur Configurer un signal d'onde arbitraire 5. EXPLORER LES FONCTIONNALITES DU GENERATEUR Dans les exemples suivants, seuls les réglages des paramètres sont présentés. Ils sont accompagnés des noms des touches correspondantes. Les formes d'onde de sortie ainsi obtenues sont affichées, ainsi que la forme d'onde de synchronisation ou de déclenchement le cas échéant. Les paramètres d'amplitude et de décalage de la sortie sont fournis à titre illustratif uniquement et ne doivent pas nécessairement être adoptés. Configurer un signal d'onde arbitraire Commencer avec l'instrument réinitialisé aux paramètres par défaut. MENU Nom de la touche programmable D'ondes Waves MENU Forme d’onde arbitraire NOM DE LA TOUCHE Arb (Arbitraire) Paramètre Sélection de l’onde Nom de la touche programmable Waves (Ondes) Réglage Sinc Fréquence - 10 kHz Amplitude Ampl 2V Décalage Offset (Décalage) 5 mV Paramètre État de sortie NOM DE LA TOUCHE Output1 (Sortie 1) Réglage On 72 TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur Configurer une forme d'onde sinusoïdale modulée AM Configurer une forme d'onde sinusoïdale modulée AM Commencer avec l'instrument réinitialisé aux paramètres par défaut. . Paramètre Nom de la touche programmable Fréquence - Réglage 10 MHz MENU Modulation NOM DE LA TOUCHE Mod Paramètre Fréquence Nom de la touche programmable - Réglage 100 kHz Profondeur Depth (Profondeur) 100 % Source Source Internal Forme Shape (Forme) Sine État de modulation On/Off On Paramètre État de sortie NOM DE LA TOUCHE Output1 (Sortie 1) Réglage On Changer la modulation et la régler sur AM-SC Paramètre Nom de la touche programmable Type Type > AM 73 Réglage AM-SC TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur PRBS PRBS Commencer avec l'instrument réinitialisé aux paramètres par défaut. MENU D'ondes NOM DE LA TOUCHE Waves MENU Bruit/PRBS Nom de la touche programmable Noise (Bruit) Réglage Source Source PBRS Paramètre Débit binaire Nom de la touche programmable BitRate Réglage 1 Mbps Amplitude Ampl 3,3 V Décalage Décalage 1,65 V Type de PRBS Type PN7 Paramètre État de sortie NOM DE LA TOUCHE Output1 (Sortie 1) Réglage On 74 TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur Modulation de fréquence d'une forme d'onde sinusoïdale Modulation de fréquence d'une forme d'onde sinusoïdale Commencer avec l'instrument réinitialisé aux paramètres par défaut. MENU Modulation NOM DE LA TOUCHE Mod Paramètre État de modulation Nom de la touche programmable On/Off Réglage On Type de modulation Type FM Fréquence de modulation - 1 kHz Déviation Deviatn 9 kHz MENU Sinusoïde NOM DE LA TOUCHE Params Paramètre Amplitude Nom de la touche programmable Ampl Réglage 1,0 V Décalage Offset (Décalage) 0,0 V Paramètre État de sortie NOM DE LA TOUCHE Output1 (Sortie 1) Réglage On 75 TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur Forme d'onde modulée à largeur d'impulsion (PWM) Forme d'onde modulée à largeur d'impulsion (PWM) Commencer avec l'instrument réinitialisé aux paramètres par défaut. MENU D'ondes NOM DE LA TOUCHE Waves Paramètre Impulsion Nom de la touche programmable Pulse MENU Modulation NOM DE LA TOUCHE Mod Paramètre État de modulation Nom de la touche programmable On/Off Réglage On Type de modulation Type PWM Fréquence de modulation - 1 kHz Déviation Dev % 40 % MENU Impulsion NOM DE LA TOUCHE Params Paramètre Amplitude Nom de la touche programmable Ampl Réglage 1,0 V Décalage Décalage 0,0 V Paramètre État de sortie NOM DE LA TOUCHE Output1 (Sortie 1) Réglage On 76 Réglage TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur Modulation par déplacement d’amplitude (ASK) Modulation par déplacement d’amplitude (ASK) Commencer avec l'instrument réinitialisé aux paramètres par défaut. MENU Modulation NOM DE LA TOUCHE Mod Paramètre Nom de la touche programmable Réglage État de modulation On/Off On Type de modulation Type ASK Source de modulation Source Internal Amplitude de saut HpAmpl 100 mV Taux de commutation Rate (Taux) 1 kHz Polarité de saut HopPol Positive MENU Sinusoïde NOM DE LA TOUCHE Params Paramètre Amplitude Nom de la touche programmable Ampl Réglage 1,0 V Décalage Décalage 0,0 V Paramètre État de sortie NOM DE LA TOUCHE Output1 (Sortie 1 ) Réglage On Notez que le deuxième tracé est la sortie de la prise Output 2 (Sortie 2) lorsque la voie 2 est synchronisée, qui suit le signal de la forme d'onde modulée. 77 TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur Balayage de fréquence d'une onde sinusoïdale Balayage de fréquence d'une onde sinusoïdale Commencer avec l'instrument réinitialisé aux paramètres par défaut. MENU Balayage NOM DE LA TOUCHE Sweep Paramètre État de balayage Nom de la touche programmable On/Off Réglage On Fréquence d’arrêt Freq > Stop 100 kHz MENU Sinusoïde NOM DE LA TOUCHE Params Paramètre Amplitude Nom de la touche programmable Ampl Réglage 1,0 V Décalage Décalage 0,0 V Paramètre État de sortie NOM DE LA TOUCHE Output1 (Sortie 1) Réglage On 78 TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur Génération d'une rafale déclenchée Génération d'une rafale déclenchée Commencer avec l'instrument réinitialisé aux paramètres par défaut. Paramètre Fréquence Nom de la touche programmable - Réglage 6 MHz MENU Rafales NOM DE LA TOUCHE Burst Paramètre État des rafales Nom de la touche programmable On/Off Réglage On Compte de rafales Count 3 Source de déclenchement SetTrg > Source > Int Internal Période de déclenchement SetTrg > Period 5 ms MENU Sinusoïde NOM DE LA TOUCHE Params Paramètre Amplitude Nom de la touche programmable Ampl Réglage 1,0 V Décalage Décalage 0,0 V Paramètre État de sortie NOM DE LA TOUCHE Output1 (Sortie 1) Réglage On Notez que le deuxième tracé est la sortie de la prise Main Output 2 (Sortie principale 2) lorsque la voie 2 est synchronisée, qui suit le signal d’entrée de déclenchement. 79 TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur Couplage des fréquences de deux voies Couplage des fréquences de deux voies Commencer avec l'instrument réinitialisé aux paramètres par défaut. MENU Utilitaires NOM DE LA TOUCHE Utility Paramètre Fréquences Nom de la touche programmable Dual Ch >Freq Réglage Coupled Paramètre État de sortie NOM DE LA TOUCHE Output1 Réglage On État de sortie Output2 On MENU Sinusoïde NOM DE LA TOUCHE Params Paramètre Changement de phase Nom de la touche programmable Output > Phase > Done Réglage 90 degrés Fréquence Freq 1MHz Veuillez noter que lorsque la voie 1 est réglée sur 1MHz, la voie 2 est également réglée sur 1MHz.Il est possible d'observer le changement de phase de 90 degrés entre les voies. 80 TGF4000 Series Quick Start Guide 5 - Explorer les fonctionnalités du générateur Compteur de fréquence Compteur de fréquence Commencer avec l'instrument réinitialisé aux paramètres par défaut. MENU Utilitaires NOM DE LA TOUCHE Utility Paramètre Compteur Nom de la touche programmable Instr > FrCntr > On/Off Réglage Counter Enabled Source Source TRIG IN - DC Coupled Type Freq Frequency Mesure Count La fenêtre Modification affiche la mesure actuelle. Sans la présence d’un signal d’entrée à l’entrée sélectionnée, le compteur affiche « No signal » jusqu'à ce qu'un signal d'entrée adéquat soit appliqué. Lorsqu’un signal d’entrée adéquat est appliqué, le compteur mesure et affiche constamment la mesure dans la case Edit. 81 TGF4000 Series Quick Start Guide 6 - Maintenance Nettoyage 6. MAINTENANCE Le fabricant ou ses agences à l’étranger proposent un service de réparation pour tout appareil défaillant. Si les propriétaires de ce matériel souhaitent en effectuer l’entretien par leurs propres moyens, il est fortement recommandé que ce travail soit effectué exclusivement par un personnel qualifié, à l’aide du guide d’entretien que l’on peut acheter directement chez le fabricant ou ses agents à l’étranger. Nettoyage Si l’appareil a besoin d’être nettoyé, utiliser uniquement un chiffon qui aura été préalablement humidifié avec un peu d’eau ou un détergent doux. AVERTISSEMENT ! AFIN D’ÉVITER TOUT RISQUE D’ÉLECTROCUTION OU UNE DÉTÉRIORATION DE L’APPAREIL, NE JAMAIS LAISSER D’EAU S’INFILTRER À L’INTÉRIEUR DU BOÎTIER. POUR ÉVITER TOUTE DÉTÉRIORATION, NE JAMAIS UTILISER DE SOLVANTS POUR NETTOYER LE BOÎTIER. 82 TGF4000 Series Quick Start Guide 0 - Inhaltsverzeichnis INHALTSVERZEICHNIS 1. Sicherheit ........................................................................................................................................................ 84 Symbole............................................................................................................................................................. 85 2. Funktionsprinzipien ...................................................................................................................................... 86 Layout des vorderen Bedienfelds ...................................................................................................................... 86 Buchsen auf der Rückseite ................................................................................................................................ 87 Bildschirmaufbau .............................................................................................................................................. 88 3. Erste Schritte ................................................................................................................................................ 89 Anfangsbedingungen......................................................................................................................................... 89 4. Grundlegende Einrichtungsbeispiele ............................................................................................................ 90 Einrichten eines Sinussignals ............................................................................................................................. 90 Einrichten eines Rechteck-Taktsignals ............................................................................................................... 94 Einrichten einer Impulswellenform ................................................................................................................. 100 Einrichten weiterer Ausgangsoptionen ........................................................................................................... 108 5. Erkunden der Generator-Funktionen .......................................................................................................... 112 Einrichten einer Arbiträr-Wellenform.............................................................................................................. 112 Einrichten einer AM-modulierten Sinuskurve ................................................................................................. 112 PRBS ................................................................................................................................................................ 114 Frequenzmodulation einer Sinuskurve............................................................................................................ 114 Pulsweitenmodulierte Wellenform (PWM) ..................................................................................................... 116 Amplitudenumtastung (ASK) ........................................................................................................................... 117 Frequenz-Sweep einer Sinuswelle ................................................................................................................... 118 Generieren eines getriggerten Burst-Signals ................................................................................................... 119 Frequenzkopplung beider Kanäle .................................................................................................................... 120 Frequenzzähler ................................................................................................................................................ 121 6. Wartung ..................................................................................................................................................... 122 Reinigung ........................................................................................................................................................ 122 Die neuesten Versionen dieses Handbuchs, Gerätetreiber und Softwaretools können von folgender Website heruntergeladen werden: www.aimtti.com 83 TGF4000 Series Quick Start Guide 1 - Sicherheit 1. SICHERHEIT Dieses Gerät wurde nach der Sicherheitsklasse (Schutzart) I der IEC-Klassifikation und gemäß den europäischen Vorschriften EN61010-1 (Sicherheitsvorschriften für Elektrische Meß-, Steuer, Regelund Laboranlagen) entwickelt. Es handelt sich um ein Gerät der Installationskategorie II, das für den Betrieb von einer normalen einphasigen Versorgung vorgesehen ist. Das Gerät wurde gemäß den Vorschriften EN61010-1 geprüft und wurde in sicherem Zustand geliefert. Die vorliegende Anleitung enthält vom Benutzer zu beachtende Informationen und Warnungen, die den sicheren Betrieb und den sicheren Zustand des Gerätes gewährleisten. Dieses Gerät ist für den Betrieb in Innenräumen der Umgebungsklass 2 , für einen Temperaturbereich von 5° C bis 40° C und 20 - 80 % relative Feuchtigkeit (nicht kondensierend) vorgesehen. Gelegentlich kann es Temperaturen zwischen +5° und –10°C ausgesetzt sein, ohne daß seine Sicherheit dadurch beeinträchtigt wird. Betreiben Sie das Gerät jedoch auf keinen Fall, solange Kondensation vorhanden ist. Ein Einsatz dieses Geräts in einer Weise, die für diese Anlage nicht vorgesehen ist, kann die vorgesehene Sicherheit beeinträchtigen. Auf keinen Fall das Gerät außerhalb der angegebenen Nennversorgungsspannungen oder Umgebungsbedingungen betreiben. WARNUNG! - DIESES GERÄT MUSS GEERDET WERDEN! Jede Unterbrechung des Netzschutzleiters innerhalb oder außerhalb des Geräts macht das Gerät gefährlich. Eine absichtliche Unterbrechung ist verboten. Die Schutzwirkung darf durch Verwendung eines Verlängerungskabels ohne Schutzleiter nicht aufgehoben werden. Ist das Gerät an die elektrische Versorgung angeschlossen, so können die Klemmen unter Spannung stehen, was bedeutet, daß beim Entfernen von Verkleidungs- oder sonstigen Teilen (mit Ausnahme der Teile, zu denen Zugang mit der Hand möglich ist) höchstwahrscheinlich spannungsführende Teile bloßgelegt weden. Vor jeglichem Öffnen des Geräts zu Nachstell-, Auswechsel-, Wartungsoder Reparaturzwecken, Gerät stets von sämtlichen Spannungsquellen abklemmen. Jegliche Nachstellung, Wartung und Reparatur am geöffneten, unter Spannung stehenden Gerät, ist nach Möglichkeit zu vermeiden. Falls unvermeidlich, sollten solche Arbeiten nur von qualifiziertem Personal ausgeführt werden, das sich der Gefahren bewußt ist. Ist das Gerät eindeutig fehlberbehaftet, bzw. wurde es mechanisch beschädigt, übermäßiger Feuchtigkeit oder chemischer Korrosion ausgesetzt, so können die Schutzeinrichtungen beeinträchtigt sein, weshalb das Gerät aus dem Verkehr zurückgezogen und zur Überprüfung und Reparatur eingesandt werden sollte. Sicherstellen, daß nur Sicherungen der vorgeschriebenen Stromstärke und des vorgesehenen Typs als Ersatz verwendet werden. Provisorische “Sicherungen” und der Kurzschluß von Sicherungshaltern ist verboten. Zur Batteriepufferung des Speichers wird bei diesem Gerät eine Lithium-Knopfzelle verwendet. Ihre typische Lebensdauer beträgt 5 Jahre. Wenn sie ausgewechselt werden muß, darf sie nur durch eine Zelle desselben Typs ersetzt werden, d.h. eine 3 V Li/MnO2 20 mm Knopfzelle, Typ 2032. Verbrauchte Zellen müssen entsprechend der am Ort geltenden Vorschriften entsorgt werden. Auf keinen Fall darf versucht werden, die Zelle aufzuladen, zu öffnen, zu verbrennen, oder sie Temperaturen von über 60° C auszusetzen. 84 TGF4000 Series Quick Start Guide 1 - Sicherheit Symbole Symbole Beim Reinigen darauf achten, daß das Gerät nicht naß wird. Am Gerät werden folgende Symbole verwendet: Vorsicht! Bitte beachten Sie die beigefügten Unterlagen. Falsche Bedienung kann Schaden am Gerät verursachen! Masse Wechselstrom Netz OFF (aus) l 85 Netz ON (ein) TGF4000 Series Quick Start Guide 2 - Funktionsprinzipien Layout des vorderen Bedienfelds 2. FUNKTIONSPRINZIPIEN Layout des vorderen Bedienfelds Ref Kurzbeschreibung Funktion 1 Stromschalter Schaltet das Gerät ein oder aus. Sicherheitshinweis: Ziehen Sie den Netzstecker an der Geräterückseite oder schalten Sie die Netzsteckdose aus, um das Gerät ganz vom Netz zu trennen. Achten Sie darauf, dass die Abschaltmöglichkeit gut zugänglich ist. 2 Softkeys Führt die Funktion auf dem LCD-Softkey aus. 3 Wellenform-Menüs Wählt den Hauptwellentyp (Trägersignal). (Sinus, Rechteck, Rampe, Puls, Rausch/PBRS oder Arb.) 4 Menüs für die WellenformModifizierung Öffnet die Menüs zum Einstellen der Parameter für Modulation, Sweep und Burst 5 Andere Menüs Wählt Menüs für interne und externe Speicherung von Dateien, Dienstprogramme und Triggerbedingungen. 6 Hauptbuchsen Hauptausgangsbuchsen. Kanal 2 kann auch so konfiguriert werden, dass das Kanal 1 Sync-Signal an der MAIN OUT 2Buchse ausgegeben wird. 7 Ausgangstasten Den gewählten MAIN OUT ein- oder ausschalten. 8 Cursor-Tasten und Drehgeber Zur Änderung numerischer Parameterwerte Ziffer für Ziffer. Zur Auswahl bestimmter Elemente innerhalb einiger Menüs. 9 Numerische Tastatur Zur direkten Eingabe numerischer Parameterwerte. 10 USB-Stick USB-Host-Anschluss für USB-Flash-Laufwerk. 86 TGF4000 Series Quick Start Guide 2 - Funktionsprinzipien Buchsen auf der Rückseite Buchsen auf der Rückseite Ref Kurzbeschreibung Funktion 1 Modulationseingang Eingang zur externen Modulation der Hauptwellenformen. 2 Referenz Eing. / ACgekoppelter Frequenzzähler Eingang für externen 10MHz Referenztakt und AC gekoppelte externe Frequenzmessung. 3 Referenz Ausg. Ausgang für internes 10 MHz Referenztaktsignal. 4 Trigger-Eingang / DCgekoppelter Frequenzzähler Eingang für externe Triggerung von Hauptwellenformen und DCgekoppelte externe Frequenzmessung 5 LAN-Schnittstelle Die LAN-Schnittstelle entspricht LXI ( Lan eXtensions for Instrumentation) Core 2011. Eine Fernsteuerung über die LANSchnittstelle ist mittels TCP/IP Socket-Protokoll möglich. 6 USB-Port Akzeptiert ein Standard-USB-Kabel. 7 GPIB (IEEE−488) -wahlweise Die Standard-GBIP-Adresse ist 5. 87 TGF4000 Series Quick Start Guide 2 - Funktionsprinzipien Bildschirmaufbau Bildschirmaufbau Ref Kurzbeschreibung Funktion 1 Kanalanzeige Zeigt momentan ausgewählten Kanal 2 Hauptwellenformtyp Zeigt das aktuelle Trägersignal 3 Ausgangsstatus Einschalten, Ausgang Ein/Aus 4 Externer Takt-Anzeige Zeigt den Status des externen Taktgebers an (falls vorhanden) 5 LAN Statusanzeige Zeigt den Status des LAN-Anschluss an (Ethernet). 6 Parameterfeld Zeigt die wichtigsten Parameter der Wellenform. 7 Menübeschreibung Zeigt das aktuelle gewählte Bearbeitungsmenü. 8 Grafikbereich Zeigt eine graphische Darstellung der ausgewählten Wellenform. 9 Bearbeitungsfeld Zeigt den aktuellen Parameter, der bearbeitet werden kann 10 Softkey-Label Zeigt die aktuellen Funktionen für die sechs Tasten unten. 88 TGF4000 Series Quick Start Guide 3 - Erste Schritte Anfangsbedingungen 3. ERSTE SCHRITTE Um den Benutzer mit einigen der grundlegenden Funktionen des Instruments vertraut zu machen, werden hier eine Reihe von Setup Beispielen beschrieben. Es wird empfohlen, dass alle Benutzer zumindest die ersten drei Beispiele ausführen sollten: • • • • Einrichten eines Sinussignals Einrichten eines Rechteck-Taktsignals Einrichten einer Impulswellenform Einrichten weiterer Ausgangsoptionen Eine Reihe weiterer Set-up Beispiele werden zur Verfügung gestellt, die eine gewisse Vertrautheit mit dem Gerät voraussetzen: • • • • • • • • • • Einrichten einer Arbiträr-Wellenform Einrichten einer AM-modulierten Sinuskurve PRBS (TGF4162 & TGF4242) Frequenzmodulation einer Sinuskurve Pulsweitenmodulierte Wellenform (PWM) Amplitudenumtastung (ASK) Frequenz-Sweep einer Sinuswelle Generieren eines getriggerten Burst-Signals Frequenzkopplung beider Kanäle Frequenzzähler Anfangsbedingungen Vor dem Einrichten des Geräts für die Beispiele sollte es auf Standardbedingungen zurückgesetzt werden. Dies geschieht folgendermaßen: • Drücken Sie die Taste mit der Bezeichnung Utility • Drücken Sie den Softkey mit der Bezeichnung System • Drücken Sie den Softkey mit der Bezeichnung Default (Display zeigt Restore Factory Default?) • Drücken Sie den Softkey Yes Dies stellt die Hauptwellenform auf Sinus (10 kHz, 1 V pk-pk) ein und stoppt alle Modulationen, Sweeps, Burst-Triggers oder Gating. ANMERKUNG: Das Gerät kann so eingestellt werden, dass es seine letzten Einstellungen beim Ausschalten speichert und diese beim Einschalten wiederherstellt. Dies wird im Utility-Menü und über den PwrOn Softkey eingestellt. Diese Einstellung geht verloren, wenn das Gerät auf die Ausgangsbedingungen zurückgesetzt wird (siehe oben). 89 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Sinussignals 4. GRUNDLEGENDE EINRICHTUNGSBEISPIELE Einrichten eines Sinussignals Voraussetzung Ausgabe eines kontinuierlichen Sinuswellensignals mit 40 MHz Frequenz und einer Amplitude von 6 V pk-pk an MAIN OUT 1. Startbedingungen Bevor Sie beginnen, sollten Sie das Gerät auf Standardwerte zurücksetzen, wie in Abschnitt Anfangsbedingungen Öffnen Sie das Wellenform-Menü - Sine (Sinus) 90 • Drücken Sie die Taste mit der Bezeichnung Waves • Drücken Sie den Softkey Sine TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Sinussignals Frequenzeinstellung Der Softkey Freq wird markiert - die aktuelle Frequenz erscheint im Eingabefeld. Beachten Sie, dass durch wiederholtes Drücken dieses Softkeys die Funktion zwischen Frequenz und Periode wechselt. • Geben Sie eine neue Frequenz mit der Nummerntastatur ein. Drücken Sie die Ziffern 4 0 Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Frequenzeinheiten an. • 91 Drücken Sie den Softkey MHz, um eine Frequenz von 40 MHz zu bestätigen. TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Sinussignals Amplitude festlegen • Drücken Sie den Softkey Ampl Beachten Sie, dass durch wiederholtes Drücken des Softkeys Ampl die Beschriftung der Tasten Ampl und Offset auf HiLvl (High-Level) und LoLvl (Low-Level) wechselt und umgekehrt. • Geben Sie eine neue Amplitude mit der Nummerntastatur ein. Drücken Sie die Ziffer 6 Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Spannungseinheiten an. 92 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Sinussignals • Drücken Sie den Softkey Vpp, um einen pk-pk Pegel von 6,0 Volt zu bestätigen. Schalten Sie den Ausgang ein • Drücken Sie die Taste Output 1 um den Kanal 1-Ausgang einzuschalten. Die Ausgangstaste 1 leuchtet orange, um den Einschaltzustand anzuzeigen 93 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Rechteck-Taktsignals Einrichten eines Rechteck-Taktsignals Voraussetzung Ausgabe eines kontinuierlichen Rechtecktaktsignals mit 20 MHz Frequenz, Tastverhältnis von 50 % und einem max. Pegel von 3,3 V und min. Pegel von 0,0 Volt an MAIN OUT 1. Startbedingungen Bevor Sie beginnen, sollten Sie das Gerät auf Standardwerte zurücksetzen, wie in Abschnitt Anfangsbedingungen Öffnen Sie das Wellenform-Menü - Square (Rechteck) 94 • Drücken Sie die Taste mit der Bezeichnung Waves • Drücken Sie den Softkey Square. TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Rechteck-Taktsignals Frequenzeinstellung Der Softkey Freq wird hervorgehoben - die aktuelle Frequenz erscheint im Eingabefeld. Beachten Sie, dass durch wiederholtes Drücken dieses Softkeys die Funktion zwischen Frequenz und Periode wechselt. • Geben Sie eine neue Frequenz mit der Nummerntastatur ein. Drücken Sie die Ziffern 2 0 Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Frequenzeinheiten an. • Drücken Sie den Softkey MHz um eine Frequenz von 20 MHz zu bestätigen. Beachten Sie, dass die Änderungen im Grafikbereich die Anstiegszeit an den Flanken zeigen, die jetzt von Bedeutung ist. 95 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Rechteck-Taktsignals Bestätigen Sie das Tastverhältnis • Drücken Sie den Softkey Duty - das aktuelle Tastverhältnis erscheint im Eingabefeld. Beachten Sie, dass das Tastverhältnis bereits auf 50 % festgelegt ist, jedoch hier ggf. geändert werden kann. Einstellen der High und Low Level • Drücken Sie den Softkey Ampl - die Tastenbeschriftung wechselt auf HiLvl und die aktuelle Spannung mit hohem Pegel erscheint im Bearbeitungsfeld Beachten Sie, dass durch wiederholtes Drücken des Softkeys Ampl die Beschriftung der Tasten Ampl und Offset auf HiLvl (High-Level) und LoLvl (Low-Level) wechselt und umgekehrt. 96 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Rechteck-Taktsignals Wenn auf dem Softkey HiLvl angegeben ist, erscheint der aktuelle High-Level Spannungspegel im Eingabefeld. • Geben Sie einen neuen Pegelwert mit der Nummerntastatur ein. Drücken Sie die Ziffern 3 . 3 Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Spannungseinheiten an. • 97 Drücken Sie den Softkey V, um einen High-Pegel von 3,3 Volt zu bestätigen. TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Rechteck-Taktsignals • Drücken Sie den Softkey LoLvl - der aktuelle Low-Level Spannungspegel erscheint im Eingabefeld. • Geben Sie einen neuen Pegelwert mit der Nummerntastatur ein. Drücken Sie 0 • Drücken Sie den Softkey V um einen Low-Pegel von 0,0 Volt zu bestätigen. Schalten Sie den Ausgang ein • Drücken Sie die Taste Output1, um den Kanal 1-Ausgang einzuschalten. Die Ausgangstaste 1 leuchtet orange, um den Einschaltzustand anzuzeigen 98 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten eines Rechteck-Taktsignals Laufende Änderungen an numerischen Parametern vornehmen (z. B. Frequenz) Numerische Parameter können mit den Cursortasten und dem Drehgeber alternativ zum numerischen Tastenfeld geändert werden. • Drücken Sie die Taste mit der Bezeichnung Waves. • Drücken Sie den Softkey Square. • Drücken Sie den Softkey Freq - der aktuelle Frequenzwert von 20,0 MHz wird angezeigt • Drücken Sie die Cursortasten, um die Bearbeitungsmarkierung auf die zweite Stelle zu setzen. • Verändern Sie den Wert mit dem Drehgeber – die Frequenz wird sofort geändert. ANMERKUNG: Drücken Sie, um das Drehgeber zu aktivieren / deaktivieren. 99 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten einer Impulswellenform Einrichten einer Impulswellenform Voraussetzung Ausgabe eines kontinuierlichen Impulssignals mit 100 ns Periode, 30 ns Pulsbreite, 20 ns Flankenzeiten, einem High Pegel von 2,7 V und einem Low Pegel von -0,6 Volt an MAIN OUT 1. Startbedingungen Bevor Sie beginnen, sollten Sie das Gerät auf Standardwerte zurücksetzen, wie in Abschnitt Anfangsbedingungen Öffnen Sie das Wellenform-Menü - Pulse (Impuls) • Drücken Sie die Taste mit der Bezeichnung Waves. • Drücken Sie den Softkey Pulse. 100 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten einer Impulswellenform Periode festlegen • Drücken Sie den Softkey PlsFrq so dass dieser auf PlsPer wechselt - die aktuelle Periode erscheint im Eingabefeld. Durch wiederholtes Drücken dieses Softkeys die Funktion zwischen Frequenz und Periode wechselt. • Geben Sie einen neuen Periodenwert mit der Nummerntastatur ein. Drücken Sie die Ziffern 1 0 0 Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Zeiteinheiten an. 101 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten einer Impulswellenform • Drücken Sie den Softkey ns, um eine Periode von 100 ns zu bestätigen. Im Grafikbereich erscheint nun eine Darstellung der Impuls- und Flankenzeiten. Impulsbreite festlegen • 102 Drücken Sie den Softkey Duty - die Tastenbeschriftung wechselt auf Width und zeigt die Breite als Zeit an TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten einer Impulswellenform • Geben Sie einen neuen Breitenwert mit der Nummerntastatur ein. Drücken Sie die Ziffern 3 0 Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Zeiteinheiten an. • 103 Drücken Sie den Softkey ns, um eine Breite von 30 ns zu bestätigen. TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten einer Impulswellenform Impulsflankenzeiten festlegen • Drücken Sie den Softkey More • Drücken Sie den Softkey Edge • Wählen Sie mit den Cursortasten die Ziffer, die Einheiten von 10 ns repräsentiert • Ändern Sie den Wert mit dem Drehgeber auf 20,0 ns Sie hätten den Wert aber auch über die numerische Tastatur eingeben können. • 104 Drücken Sie den Softkey Done. TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten einer Impulswellenform Einstellen der High und Low Level • Drücken Sie den Softkey Ampl - die Tastenbeschriftung wechselt auf HiLvl und die aktuelle Spannung mit hohem Pegel erscheint im Bearbeitungsfeld Beachten Sie, dass durch wiederholtes Drücken des Softkeys Ampl die Beschriftung der Tasten Ampl und Offset auf HiLvl (High-Level) und LoLvl (Low-Level) wechselt und umgekehrt. • Geben Sie einen neuen Pegelwert mit der Nummerntastatur ein. Drücken Sie die Ziffern 2 . 7. Sobald eine Zahl eingegeben wird, ändern sich die Softkeys und zeigen Spannungseinheiten an. 105 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten einer Impulswellenform • Drücken Sie den Softkey V, um einen High-Pegel von 2,7 Volt zu bestätigen. • Drücken Sie den Softkey LoLvl - der aktuelle Low-Level Spannungspegel erscheint im Eingabefeld. 106 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten einer Impulswellenform • Geben Sie einen neuen Pegelwert mit der Nummerntastatur ein. Drücken Sie - . 6 • Drücken Sie den Softkey V, um einen Low-Pegel von -600 mV zu bestätigen. Schalten Sie den Ausgang ein • Drücken Sie die Taste Output1, um den Kanal 1-Ausgang einzuschalten. Die Ausgangstaste 1 leuchtet orange, um den Einschaltzustand anzuzeigen 107 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten weiterer Ausgangsoptionen Einrichten weiterer Ausgangsoptionen Voraussetzung In den früheren Beispielen wurde gezeigt, wie das Ausgangsmenü verwendet wird, um den Ausgangspegel einzustellen (Amplitude plus Offset oder High Pegel plus Low Pegel) und den Ausgang ein- oder auszuschalten. Dieses Beispiel zeigt die Einstellung der Ausgangsphase, Ausgangspolarität, Lastimpedanz und Spannungsautomatik. Startbedingungen Bevor Sie beginnen, sollten Sie das Gerät auf Standardwerte zurücksetzen, wie in Abschnitt Anfangsbedingungen Öffnen Sie das Ausgangsmenü • Drücken Sie den Softkey Output 108 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten weiterer Ausgangsoptionen Ändern der Ausgangsphase • Standardmäßig ist der Softkey Phase ausgewählt. • Geben Sie eine Phase von - 4 5 Grad ein. Der eingestellte Phasenwinkel ist der Punkt in der Wellenformperiode, der mit dem Sync oder der Triggerflanke übereinstimmt, d. h. es ist der Punkt in der Periode, an der die Wellenform beginnt. Daher eilt eine negative Phaseneinstellung voraus, während eine positive Phaseneinstellung die Wellenform in Bezug auf den Sync oder Trigger verzögert. Dies wird auch im Grafikbereich angezeigt. Der Softkey ‚Align’ dient zur Neuausrichtung der Phase bei Frequenzänderungen. Ändern der Ausgangspolarität • Drücken Sie den Softkey Type, um die Ausgangspolarität zu invertieren. Beachten Sie, dass wiederholtes Drücken der Taste Type einen Wechsel zwischen normal und invertiert bewirkt. 109 TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten weiterer Ausgangsoptionen Ändern der Lastimpedanz • Drücken Sie den Softkey Load • Drücken Sie den Softkey Load erneut, um die Lastimpedanz auf „High-z“ (hohe Impedanz) zu ändern. Beachten Sie, dass die Standard-Lastimpedanz 50 Ohm beträgt, jedoch auf eine beliebige Impedanz zwischen 50 und 10.000 Ohm eingestellt werden kann. Die Pegel werden auf der Grundlage dieser Impedanz berechnet. Beachten Sie, dass wiederholtes Drücken der Taste Load einen Wechsel zwischen einem numerischen Wert und „High-z“ bewirkt. Beachten Sie, dass sich der angezeigte Amplitudenwert auf 2 Volt pk-pk erhöht. • 110 Drücken Sie den Softkey Load, um die Lastimpedanz wieder auf 50 Ohm zu setzen. TGF4000 Series Quick Start Guide 4 - Grundlegende Einrichtungsbeispiele Einrichten weiterer Ausgangsoptionen Bereichsumschaltung • Drücken Sie den Softkey Range • Drücken Sie den Softkey Range erneut, um den Bereich von Auto auf Hold zu ändern. Die Spannungsautomatik verläuft in Dämpfungsschritten von jeweils 6dB (d. h. ‘Range‘ Maximalwerte von 10 Vpp, 5 Vpp, 2,5 Vpp etc. an 50 Ω), wobei die Feineinstellung der Amplitude auf 6dB begrenzt ist, um die Güte der Wellenform zu erhalten. Durch Wahl von Hold wird die automatische Bereichswahl deaktiviert. Die Dämpfungseinstellung ist fixiert und die Feineinstellung der Amplitude nicht länger begrenzt. Wenn ‚Range’ auf Auto eingestellt ist, schalten Amplitude und Abschwächer automatisch, sodass eine optimale Leistung gegeben ist. Bei Bereichseinstellung ‚Hold‘ wird für alle Amplitudeneinstellungen eine feste Dämpfungseinstellung verwendet. 111 TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen Einrichten einer Arbiträr-Wellenform 5. ERKUNDEN DER GENERATOR-FUNKTIONEN In den folgenden Beispielen werden nur die Parametereinstellungen beschrieben, zusammen mit den dazugehörigen Tastenbezeichnungen. Es werden zudem die resultierenden Ausgangswellenformen gezeigt (zusammen mit der Sync- oder Triggerwellenform, falls relevant). Ausgangsamplitude und Offset-Einstellungen sind nur Beispiele und müssen nicht beachtet werden. Einrichten einer Arbiträr-Wellenform Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. MENÜ TASTENBEZEICHNUNG Wellenform Waves MENÜ Arbiträr-Wellenform Softkey-Bezeichnung Arb Parameter Kurvenauswahl Softkey-Bezeichnung Waves Einstellung Sinc Frequenz - 10 kHz Amplitude Ampl 2V Offset Offset 5mV Parameter Ausgangsstatus TASTENBEZEICHNUNG Output1 Einstellung On 112 TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen Einrichten einer AM-modulierten Sinuskurve Einrichten einer AM-modulierten Sinuskurve Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. Parameter Softkey-Bezeichnung Frequenz - Einstellung 10MHz MENÜ Modulation TASTENBEZEICHNUNG Mod Parameter Frequenz Softkey-Bezeichnung - Einstellung 100kHz Tiefe Depth 100% Quelle Source Internal Form Shape Sinus Modulationstatus On/Off On Parameter Ausgangsstatus TASTENBEZEICHNUNG Output1 Einstellung On Modulation auf AM-SC ändern Parameter Typ 113 Softkey-Bezeichnung Type > AM Einstellung AM-SC TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen PRBS PRBS Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. MENÜ Wellenform TASTENBEZEICHNUNG Waves MENÜ Rauschen/PRBS Softkey-Bezeichnung Noise Einstellung Quelle Source PBRS Parameter Bit-Rate Softkey-Bezeichnung BitRate Einstellung 1Mbps Amplitude Ampl 3.3V Offset Offset 1.65V PRBS Typ Type PN7 Parameter Ausgangsstatus TASTENBEZEICHNUNG Output1 Einstellung On 114 TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen Frequenzmodulation einer Sinuskurve Frequenzmodulation einer Sinuskurve Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. MENÜ Modulation TASTENBEZEICHNUNG Mod Parameter Modulationsstatus Softkey-Bezeichnung On/Off Einstellung On Modulationstyp Type FM Modulationsfrequenz - 1kHz Abweichung Deviatn 9kHz MENÜ Sinus TASTENBEZEICHNUNG Params Parameter Amplitude Softkey-Bezeichnung Ampl Einstellung 1.0V Offset Offset 0.0V Parameter Ausgangsstatus TASTENBEZEICHNUNG Output1 Einstellung On 115 TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen Pulsweitenmodulierte Wellenform (PWM) Pulsweitenmodulierte Wellenform (PWM) Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. MENÜ Wellenform TASTENBEZEICHNUNG Waves Parameter Puls Softkey-Bezeichnung Pulse MENÜ TASTENBEZEICHNUNG Modulation Mod Parameter Modulationsstatus Softkey-Bezeichnung On/Off Einstellung On Modulationstyp Type PWM Modulationsfrequenz - 1kHz Abweichung Dev % 40% MENÜ Pulse TASTENBEZEICHNUNG Params Parameter Amplitude Softkey-Bezeichnung Ampl Einstellung 1.0V Offset Offset 0.0V Parameter Ausgangsstatus TASTENBEZEICHNUNG Output1 Einstellung On 116 TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen Amplitudenumtastung (ASK) Amplitudenumtastung (ASK) Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. MENÜ Modulation TASTENBEZEICHNUNG Mod Parameter Modulationsstatus Softkey-Bezeichnung On/Off Einstellung On Modulationstyp Type ASK Modulationsquelle Source Internal Hop-Amplitude HpAmpl 100mV Umtastrate Rate 1kHz Hop-Polarität HopPol Positive MENÜ Sinus TASTENBEZEICHNUNG Params Parameter Amplitude Softkey-Bezeichnung Ampl Einstellung 1.0V Offset Offset 0.0V Parameter Ausgangsstatus TASTENBEZEICHNUNG Output1 Einstellung On Beachten Sie, dass die zweite Messkurve das Ausgangssignal der Hauptbuchse von Ausgang 2 darstellt, wenn Kanal 2 synchronisiert ist und dem modulierenden Wellenformsignal folgt. 117 TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen Frequenz-Sweep einer Sinuswelle Frequenz-Sweep einer Sinuswelle Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. MENÜ Sweep TASTENBEZEICHNUNG Sweep Parameter Sweep-Status Softkey-Bezeichnung On/Off Einstellung On Stoppfrequenz Freq > Stop 100kHz MENÜ Sinus TASTENBEZEICHNUNG Params Parameter Amplitude Softkey-Bezeichnung Ampl Einstellung 1.0V Offset Offset 0.0V Parameter Ausgangsstatus TASTENBEZEICHNUNG Output1 Einstellung On 118 TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen Generieren eines getriggerten Burst-Signals Generieren eines getriggerten Burst-Signals Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. Parameter Frequenz Softkey-Bezeichnung - Einstellung 6MHz MENÜ Burst TASTENBEZEICHNUNG Burst Parameter Burst-Status Softkey-Bezeichnung On/Off Einstellung On Burst-Count Count 3 Triggerquelle SetTrg > Source > Int Internal Trigger Triggerperiode SetTrg > Period 5ms MENÜ Sinus TASTENBEZEICHNUNG Params Parameter Amplitude Softkey-Bezeichnung Ampl Einstellung 1.0V Offset Offset 0.0V Parameter Ausgangsstatus TASTENBEZEICHNUNG Output1 Einstellung On Beachten Sie, dass die zweite Messkurve das Ausgangssignal der Hauptbuchse von Ausgang 2 darstellt, wenn Kanal 2 synchronisiert ist und dem modulierenden Trigger-Eingangssignal folgt. 119 TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen Frequenzkopplung beider Kanäle Frequenzkopplung beider Kanäle Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. MENÜ Utility TASTENBEZEICHNUNG Utility Parameter Frequenzen Softkey-Bezeichnung Dual Ch >Freq Einstellung Coupled Parameter Ausgangsstatus TASTENBEZEICHNUNG Output1 Einstellung On Ausgangsstatus Output2 On MENÜ Sinus TASTENBEZEICHNUNG Params Parameter Phasenverschiebung Softkey-Bezeichnung Output > Phase > Done Einstellung 90 Degrees Frequenz Freq 1MHz Beachten Sie, dass, wenn Kanal 1 auf 1MHz eingestellt wird, Kanal 2 ebenfalls auf 1MHz eingestellt ist. Die 90-Grad-Phasenverschiebung zwischen den Kanälen ist zu erkennen. 120 TGF4000 Series Quick Start Guide 5 - Erkunden der Generator-Funktionen Frequenzzähler Frequenzzähler Setzen Sie das Gerät zunächst auf die Standardeinstellungen zurück. MENÜ Utility TASTENBEZEICHNUNG Utility Parameter Zähler Softkey-Bezeichnung Instr > FrCntr > On/Off Einstellung Counter Enabled Quelle Source TRIG IN - DC-Coupled Typ Freq Frequency Messung Count Der Editierbereich zeigt die aktuelle Messung. Wenn am ausgewählten Eingang kein Eingangssignal anliegt, zeigt der Zähler ‚No signal‘ an, bis ein entsprechendes Eingangssignal anliegt. Wenn ein ausreichendes Eingangssignal anliegt, misst der Zähler ständig und zeigt den Messwert im Editierbereich an. 121 TGF4000 Series Quick Start Guide 6 - Wartung Reinigung 6. WARTUNG Die Hersteller oder ihre Vertretungen bieten einen Reparaturdienst für fehlerhafte Geräte an. Falls Anwender Wartungsarbeiten selbst durchführen möchten, sollten sie nur geschultes Personal damit beauftragen. Für diese Arbeiten sollte das Servicehandbuch zu Hilfe genommen werden, das direkt beim Hersteller der Geräte oder dessen Vertretungen bezogen werden kann. Reinigung Verwenden Sie zur Reinigung des Geräts ein leicht mit Wasser oder einem milden Reinigungsmittel angefeuchtetes Tuch. ACHTUNG! ZUR VERMEIDUNG VON STROMSCHLÄGEN ODER BESCHÄDIGUNGEN DES GERÄTS DARF KEIN WASSER IN DAS GEHÄUSE GELANGEN. DAS GERÄT NICHT MIT LÖSUNGSMITTELN REINIGEN, UM SCHÄDEN AM GEHÄUSE ZU VERMEIDEN. 122 TGF4000 Series Quick Start Guide 0 - Indice INDICE 1. Sicurezza 124 Simboli 125 2. Principi di funzionamento 126 Layout del pannello anteriore 126 Layout del pannello posteriore 127 Layout dello schermo 128 3. Per iniziare 129 Condizioni iniziali 129 4. Esempi di configurazione di base 130 Configurazione di un segnale a onda sinusoidale 130 Configurazione di un segnale di clock di onda quadrata 134 Configurazione di una forma d’onda a impulso 140 Configurazione di altre opzioni di uscita 148 5. Analisi delle funzionalità del generatore 152 Configurazione di un segnale d’onda arbitrario 152 Configurazione di una forma d’onda sinusoidale modulata AM 153 PRBS 154 Modulazione di frequenza di una forma d’onda sinusoidale 154 Forma d’onda modulata ad ampiezza di impulso (PWM) 156 Amplitude Shift Keying (ASK) (Modulazione a spostamento di ampiezza) 157 Frequenza di scansione di un’onda sinusoidale 158 Generazione di un burst triggered 158 Accoppiamento della frequenza di entrambi i canali 160 Contatore di frequenza 161 6. Manutenzione 162 Pulizia 162 Le ultime revisioni di questo manuale, driver di dispositivo e strumenti software possono essere scaricati da: www.aimtti.com 123 TGF4000 Series Quick Start Guide 1 - Sicurezza 1. SICUREZZA Il generatore è uno strumento di sicurezza classe I in base alla classificazione IEC ed è stato studiato per corrispondere ai requisiti EN61010-1 (requisiti sulla sicurezza di attrezzature elettriche di misurazione, controllo e uso in laboratorio). E’ uno strumento di installazione categoria II destinato al funzionamento attraverso alimentazione a fase singola normale. Lo strumento è stato testato in conformità a EN61010-1 ed è stato fornito in condizioni di sicurezza. Il presente manuale di istruzioni comprende alcune informazioni e avvertenze che devono essere seguite dall’utente al fine di garantire un funzionamento sicuro e la conservazione dello strumento in condizione di sicurezza. Lo strumento è stato studiato per l’uso all’interno in ambiente con grado di inquinamento 2 con range di temperatura da 5 °C a 40 °C, 20%-80% UR (senza formazione di condensa). Può essere occasionalmente esposto a temperature comprese fra +5 °C e –10 °C senza compromissione della sicurezza. Non utilizzare in presenza di condensa. L’uso dello strumento in modo non specificato dalle presenti istruzioni può compromettere la protezione di sicurezza fornita. Non utilizzare lo strumento al di fuori del tensione o delle condizioni ambientali indicate. AVVERTENZA! LO STRUMENTO DEVE PROVVISTO DI MESSA A TERRA Qualsiasi interruzione del conduttore di terra all’interno o all’esterno dello strumento rende pericoloso l’uso dello strumento. Un’interruzione intenzionale è vietata. Non compromettere la sicurezza mediante l’uso di una prolunga senza conduttore di protezione. Quando lo strumento è collegato all’alimentazione, i terminali possono essere scoperti e l’apertura delle coperture o la rimozione di parti (eccetto quelle accessibili con le mani) possono esporre parti esposte. Prima dell’apertura per regolazioni, sostituzioni, manutenzione o riparazione, lo strumento deve essere scollegato da tutte le sorgenti di alimentazione. Qualsiasi regolazione, manutenzione o riparazione dello strumento aperto in tensione deve essere evitata e se inevitabile, deve essere effettuata esclusivamente da personale competente, consapevole del possibile pericolo. In caso di evidente difetto dello strumento, danno meccanico, umidità eccessiva o corrosione chimica, la protezione di sicurezza potrebbe essere compromessa e il dispositivo non dovrà più essere utilizzato e sottoposto a controllo e riparazione. Accertare di utilizzare solo i fusibili della potenza e del tipo prescritti per eventuali sostituzioni. L’uso di fusibili improvvisati e la cortocircuitazione del portafusibili è vietato. Lo strumento utilizza una pila a bottone al litio per alimentare la memoria di sicurezza non volatile; la pila dura in media 5 anni. Qualora sia necessario sostituirla, utilizzare una pila del tipo corretto, ovvero una pila a bottone da 3 V Li/Mn02 da 20 mm tipo 2032. Smaltire le pile usate osservando le disposizioni vigenti; non aprire, tagliare, bruciare od esporre la pila a temperature superiori ai 60°C e non sottoporla a ricarica. Non bagnare lo strumento durante la pulizia e utilizzare esclusivamente un panno asciutto per pulire lo schermo LCD. 124 TGF4000 Series Quick Start Guide 1 - Sicurezza Simboli Simboli I seguenti simboli vengono utilizzati sullo strumento e nel manuale: Attenzione - vedere i documenti allegati. L’uso errato può danneggiare lo strumento terminale collegato al fondo dello chassis alimentazione principale OFF. l alimentazione principale ON. corrente alternata. 125 TGF4000 Series Quick Start Guide 2 - Principi di funzionamento Layout del pannello anteriore 2. PRINCIPI DI FUNZIONAMENTO Layout del pannello anteriore Rif Descrizione breve Funzione 1 Interruttore di alimentazione Accende e spegne lo strumento. Nota di sicurezza: Per scollegare completamente l'alimentazione CA, staccare il cavo di alimentazione dal retro dello strumento oppure scollegare dalla presa di alimentazione CA. Assicurarsi che i punti di scollegamento siano facilmente accessibili. 2 Soft-key: Esegue la funzione mostrata nell’etichetta del soft-key dell’LCD precedente. 3 Menu Waveform Seleziona il tipo di forma d’onda principale (forma d’onda del vettore) come attivo. (sinusoidi, quadre, a dente di sega, a impulso, di rumore/PBRS e arbitrarie). 4 Menu Waveform Modification Apre il menu per l’impostazione dei parametri di modulazione, scansione e burst 5 Altri menu Seleziona i menu per l’archiviazione interna ed esterna dei file, le utility degli strumenti e le condizioni di attivazione. 6 Socket principali Socket di uscita principali. Il canale 2 può anche essere configurato per la sincronizzazione con il canale di uscita 1 dal socket MAIN OUT 2. 7 Tasti uscita Posizionare la MAIN OUT selezionata su On o Off. 8 Tasti cursore e rotellina Utilizzato per cambiare i valori numerici dei parametri cifra per cifra. Utilizzato per selezionare le voci all’interno di alcuni menu. 9 Tastierino numerico Utilizzato per inserire direttamente i valori numerici dei parametri. 10 Unità flash USB Connettore dell’host USB per archiviazione nell’unità flash USB. 126 TGF4000 Series Quick Start Guide 2 - Principi di funzionamento Layout del pannello posteriore Layout del pannello posteriore Rif Descrizione breve Funzione 1 Ingresso modulazione Ingresso esterno per modulazione delle forme d’onda principali. 2 Ingresso di riferimento/Contatore di frequenza accoppiato CA Ingresso del clock di riferimento esterno da 10MHz e misurazione della frequenza esterna accoppiata CA. 3 Uscita di riferimento Uscita per un clock interno di riferimento da 10 MHz. 4 Ingresso di generazione/Contatore di frequenza accoppiato CC Ingresso esterno per la generazione delle forme d’onda principali e misurazione della frequenza esterna accoppiata CC 5 LAN L'interfaccia LAN è stata progettata per soddisfare i requisiti LXI Core 2011. Il protocollo TCP/IP Socket consente di attivare il controllo a distanza tramite l'interfaccia LAN. 6 USB Questa accetta cavi USB standard 7 GPIB (IEEE-488) - opzionale L'indirizzo GPIB predefinito è 5. 127 TGF4000 Series Quick Start Guide 2 - Principi di funzionamento Layout dello schermo Layout dello schermo Rif Descrizione breve Funzione 1 Indicatore di canale Mostra il canale attualmente selezionato 2 Tipo di forma d’onda principale Mostra l’attuale forma d’onda del vettore 3 Stato uscita Indica se l’uscita principale è attivata o disattivata 4 Indicatore del clock esterno. Mostra lo stato del clock esterno (se applicato) 5 Indicatore stato LAN Mostra lo stato della connessione LAN (Ethernet). 6 Casella dei parametri Mostra i parametri principali della forma d’onda. 7 Descrizione del menu Mostra il menu di modifica attualmente selezionato. 8 Casella grafica Mostra una rappresentazione grafica della forma d’onda selezionata. 9 Casella di modifica Mostra il parametro corrente che può essere modificato 10 Etichette soft-key Mostra le funzioni correnti dei sei tasti seguenti. 128 TGF4000 Series Quick Start Guide 3 - Per iniziare Condizioni iniziali 3. PER INIZIARE Per familiarizzare l’utente con alcune delle funzionalità di base dello strumento, in questa guida vengono mostrati alcuni esempi di configurazione. Si consiglia a tutti gli utenti di applicare i primi tre esempi: • • • • Configurazione di un segnale a onda sinusoidale Configurazione di un segnale di clock di onda quadrata Configurazione di una forma d’onda a impulso Configurazione di altre opzioni di uscita Vengono forniti alcuni esempi di configurazione che presuppongono una maggiore familiarità con lo strumento: • • • • • • • • • • Configurazione di un segnale d’onda arbitrario Configurazione di una forma d’onda sinusoidale modulata AM PRBS (TGF4162 & TGF4242) Modulazione di frequenza di una forma d’onda sinusoidale Forma d’onda modulata ad ampiezza di impulso (PWM) Amplitude Shift Keying (ASK) (Modulazione a spostamento di ampiezza) Frequenza di scansione di un’onda sinusoidale Generazione di un burst triggered Accoppiamento della frequenza di entrambi i canali Contatore di frequenza Condizioni iniziali Prima di configurare lo strumento per uno degli esempi, deve essere ripristinato alle condizioni predefinite. A questo scopo, procedere come segue: • • • • Premere il tasto fisso contrassegnato con Utility. Premere il soft-key etichettato System Premere il soft-key etichettato Default (lo schermo indica Restore Factory Default?) Premere il soft-key etichettato Yes La forma d’onda principale viene impostata Sinusoide (10kHz, 1V picco-picco) e vengono annullate le modulazioni, le scansioni o il burst triggering o gating. NOTA: Lo strumento può essere impostato per conservare le impostazioni più recenti allo spegnimento e ripristinarle all’accensione. Questa opzione viene impostata dal menu Utility e dal soft-key PwrOn. Questa impostazione va perduta quando vengono ripristinate le condizioni predefinite dello strumento, come sopra descritto. 129 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale a onda sinusoidale 4. ESEMPI DI CONFIGURAZIONE DI BASE Configurazione di un segnale a onda sinusoidale Requisito Emettere un segnale continuo a onda sinusoidale con una frequenza di 40MHz e un’ampiezza di 6 volt picco-picco da MAIN OUT 1. Condizioni di avvio Prima dell’avvio, ripristinare lo strumento alle impostazioni predefinite, come descritto nella sezione Condizioni iniziali Apertura del menu della forma d’onda - Sinusoidale • Premere il tasto fisso contrassegnato con Waves • Premere il soft-key etichettato Sine 130 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale a onda sinusoidale Impostazione della frequenza Il soft-key etichettato Freq viene evidenziato; la frequenza corrente viene mostrata nella casella di modifica. Tenere presente che premendo più volte il soft-key la funzione cambia tra Frequency e Period. • Utilizzare il tastierino numerico per inserire una nuova frequenza. Premere i numeri 4 0 Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di frequenza. • 131 Premere il soft-key etichettato MHz per confermare una frequenza di 40MHz. TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale a onda sinusoidale Impostazione dell’ampiezza • Premere il soft-key etichettato Ampl Le successive pressioni del soft-key Ampl cambiano le etichette dei tasti Ampl e Offset in HiLvl (livello alto) e LoLvl (livello basso) e viceversa. • Utilizzare il tastierino numerico per inserire una nuova ampiezza. Premere il numero 6 Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tensione. 132 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale a onda sinusoidale • Premere il soft-key etichettato Vpp per confermare un’ampiezza picco-picco di 6,0 volt. Attivazione dell’uscita • Premere il tasto Output 1 per attivare l’uscita del canale 1. Il tasto di uscita 1 si illumina in arancione per indicare lo stato attivo 133 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale di clock di onda quadrata Configurazione di un segnale di clock di onda quadrata Requisito Emettere un segnale continuo di clock a onda quadrata con una frequenza di 20MHz, 50% ciclo di funzionamento, un livello elevato di 3,3V e un livello basso di 0,0 volt da MAIN OUT 1. Condizioni di avvio Prima dell’avvio, ripristinare lo strumento alle impostazioni predefinite, come descritto nella Condizioni iniziali Apertura del menu della forma d’onda - Quadrata • Premere il tasto fisso contrassegnato con Waves • Premere il soft-key etichettato Square. 134 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale di clock di onda quadrata Impostazione della frequenza Il soft-key etichettato Freq viene evidenziato; la frequenza corrente viene mostrata nella casella di modifica. Tenere presente che premendo più volte il soft-key la funzione cambia tra Frequency e Period. • Utilizzare il tastierino numerico per inserire una nuova frequenza. Premere i numeri 2 0. Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di frequenza. • Premere il soft-key etichettato MHz per confermare una frequenza di 20MHz. Tenere presente che la casella grafica cambia e mostra il tempo di salita ai fronti, che adesso è un valore significativo. 135 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale di clock di onda quadrata Conferma del ciclo di funzionamento • Premere il soft-key etichettato Duty; il ciclo di funzionamento corrente viene mostrato nella casella di modifica. Tenere presente che il ciclo di funzionamento è già impostato sul 50%, ma può essere modificato se necessario. Impostazione del livello alto e del livello basso • Premere il soft-key etichettato Ampl; l’etichetta del tasto cambia in HiLvl e la tensione del livello alto corrente viene mostrata nella casella di modifica Le successive pressioni del soft-key Ampl cambiano le etichette dei tasti Ampl e Offset in HiLvl (livello alto) e LoLvl (livello basso) e viceversa. 136 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale di clock di onda quadrata Quando il soft-key è etichettato HiLvl, la tensione del livello alto corrente viene mostrata nella casella di modifica. • Utilizzare il tastierino numerico per inserire un nuovo livello. Premere i numeri 3 . 3 Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tensione. • 137 Premere il soft-key etichettato V per confermare un livello alto di 3.3 volt. TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale di clock di onda quadrata • Premere il soft-key etichettato LoLvl; la tensione del livello basso corrente viene mostrata nella casella di modifica. • Utilizzare il tastierino numerico per inserire un nuovo livello. Premere 0 . • Premere il soft-key etichettato V per confermare un livello basso di 0,0 volt. Attivazione dell’uscita • Premere il tasto Output1 per attivare l’uscita del canale 1. Il tasto di uscita 1 si illumina in arancione per indicare lo stato attivo 138 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di un segnale di clock di onda quadrata Modifiche in tempo reale a un parametro numerico (es. frequenza) I parametri numerici possono essere modificati utilizzando i tasti cursore e la rotellina in alternativa al tastierino numerico. • Premere il tasto fisso contrassegnato con Waves • Premere il soft-key etichettato Square. • Premere il soft-key etichettato Freq; viene visualizzato il valore corrente della frequenza di 20,0MHz. • Premere i tasti fissi Cursor per spostare la selezione della modifica alla seconda cifra. • Usare la rotellina per modificare il valore; la frequenza viene immediatamente modificata. NOTA: Premere per attivare / disattivare la rotellina. 139 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di una forma d’onda a impulso Configurazione di una forma d’onda a impulso Requisito Emettere un segnale continuo a impulsi con un periodo di 100ns, ampiezza di 30ns, durata dei fronti di 20ns e un livello alto di 2,7V e un livello basso di -0,6 volt da MAIN OUT 1. Condizioni di avvio Prima dell’avvio, ripristinare lo strumento alle impostazioni predefinite, come descritto nella Condizioni iniziali Apertura del menu della forma d’onda - Impulso • Premere il tasto fisso contrassegnato con Waves • Premere il soft-key etichettato Pulse. 140 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di una forma d’onda a impulso Impostazione del periodo • Premere il soft-key etichettato PlsFrq in modo che cambi in PlsPer; viene visualizzato il periodo corrente nella casella di modifica. Tenere presente che premendo più volte il soft-key la funzione cambia tra Frequency e Period. • Utilizzare il tastierino numerico per inserire un nuovo periodo. Premere i numeri 1 0 0 Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tempo. 141 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di una forma d’onda a impulso • Premere il soft-key etichettato ns per confermare un periodo di 100ns. Notare che la casella grafica cambia per mostrare una rappresentazione della durata dell’impulso e dei fronti. Impostazione dell’ampiezza dell’impulso • Premere il soft-key etichettato Duty; l’etichetta cambia in Width e mostra la larghezza come unità di tempo 142 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di una forma d’onda a impulso • Utilizzare il tastierino numerico per inserire una nuova ampiezza. Premere i numeri 3 0 Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tempo. • 143 Premere il soft-key etichettato ns per confermare un’ampiezza di 30ns. TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di una forma d’onda a impulso Impostazione della durata dei fronti dell’impulso • Premere il soft-key etichettato More • Premere il soft-key etichettato Edge • Usare i tasti cursore per selezionare la cifra che rappresenta unità di 10ns • Usare la rotellina per cambiare il valore in 20,0ns Se si preferisce, il valore può essere inserito con il tastierino numerico. • 144 Premere il soft-key etichettato Done. TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di una forma d’onda a impulso Impostazione del livello alto e del livello basso • Premere il soft-key etichettato Ampl; l’etichetta del tasto cambia in HiLvl e la tensione del livello alto corrente viene mostrata nella casella di modifica Le successive pressioni del soft-key Ampl cambiano le etichette dei tasti Ampl e Offset in HiLvl (livello alto) e LoLvl (livello basso) e viceversa. • Utilizzare il tastierino numerico per inserire un nuovo livello. Premere i numeri 2 . 7 . Non appena viene inserito il numero, i soft-key cambiano per mostrare le unità di tensione. 145 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di una forma d’onda a impulso • Premere il soft-key etichettato V per confermare un livello alto di 2,7 volt. • Premere il soft-key etichettato LoLvl; la tensione del livello basso corrente viene mostrata nella casella di modifica. 146 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di una forma d’onda a impulso • Utilizzare il tastierino numerico per inserire un nuovo livello. Premere - . 6 • Premere il soft-key etichettato V per confermare un livello basso di -600 mV. Attivazione dell’uscita • Premere il tasto Output1 per attivare l’uscita del canale 1. Il tasto di uscita 1 si illumina in arancione per indicare lo stato attivo 147 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di altre opzioni di uscita Configurazione di altre opzioni di uscita Requisito Nei precedenti esempi di configurazione si è mostrato come usare il menu Output per impostare il livello di uscita (ampiezza più offset o livello alto più livello basso) e attivare o disattivare l’uscita. Questo esempio mostra l’impostazione della fase di uscita, la polarità dell’uscita, l’impedenza di carico e la calibrazione automatica della tensione. Condizioni di avvio Prima dell’avvio, ripristinare lo strumento alle impostazioni predefinite, come descritto nella Condizioni iniziali Aprire il menu Output • 148 Premere il soft-key contrassegnato con Output TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di altre opzioni di uscita Modifica della fase di uscita Il soft-key etichettato come Phase viene selezionato per impostazione predefinita. • Inserire una fase di -45 gradi. L’angolo di fase impostato è il punto nel periodo della forma d’onda che coincide con il fronte di sincronizzazione o di generazione, ovvero il punto in cui si origina la forma d’onda. Quindi, un’impostazione negativa della fase anticipa la forma d’onda in relazione alla sincronizzazione o alla generazione, mentre un’impostazione positiva la ritarda. La forma d’onda nella casella grafica cambia per riflettere questo fenomeno. Il soft-key Align serve per riallineare la fase quando si modifica la frequenza. Modifica della polarità dell’uscita • Premere il soft-key etichettato Type per invertire la polarità dell’uscita. Le successive pressioni del tasto Type alternano tra polarità normale e invertita. 149 TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di altre opzioni di uscita Modifica dell’impedenza di carico • Premere il soft-key etichettato Load • Premere nuovamente il soft-key etichettato Load per cambiare l’impedenza di carico in High-z (alta impedenza). L’impedenza di carico predefinita è 50 Ohm, ma può essere modificata in qualsiasi valore compreso tra 50 e 10.000 Ohm. I livelli sono calcolati in base a questa impedenza. Le successive pressioni del tasto Load alternano tra un valore numerico e High-z. Tenere presente che il valore dell’ampiezza aumenta a 2 volt picco-picco. • 150 Premere il soft-key etichettato Load per tornare all’impedenza di carico di 50 Omh. TGF4000 Series Quick Start Guide 4 - Esempi di configurazione di base Configurazione di altre opzioni di uscita Modifica dell’intervallo • Premere il soft-key etichettato Range • Premere nuovamente il soft-key etichettato Range per cambiare l’intervallo da Auto a Hold. La modalità Auto esegue automaticamente la calibrazione in incrementi dell’attenuatore da 6dB (ovvero, valori massimi dell’intervallo di 10Vpp, 5Vpp, 2,5Vpp ecc. in 50Ω), con l’intervallo Vernier dell’ampiezza limitato a 6dB per mantenere la qualità della forma d’onda. Selezionando la modalità Hold viene disattivata la calibrazione automatica; l’impostazione dell’attenuatore è fissa e l’intervallo Vernier dell’ampiezza non è più limitato. Con l'intervallo impostato su Auto l'ampiezza e gli attenuatori vengono cambiati automaticamente e si ottiene la performance ottimale. Con l’intervallo impostato su Hold viene utilizzata un’impostazione fissa dell’attenuatore per tutte le impostazioni dell’ampiezza. 151 TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore Configurazione di un segnale d’onda arbitrario 5. ANALISI DELLE FUNZIONALITÀ DEL GENERATORE Nei seguenti esempi vengono descritte solo le impostazioni dei parametri, insieme ai nomi dei tasti correlati. Vengono mostrate le forme d’onda dell’uscita risultanti, insieme alla forma d’onda di sincronizzazione o generazione, dove pertinente. Le impostazioni di ampiezza dell’uscita e offset sono solo esempi, che non è obbligatorio applicare. Configurazione di un segnale d’onda arbitrario Iniziare con lo strumento configurato con le impostazioni predefinite. MENU NOME DEL TASTO FISSO Forma d'onda arbitraria Waves MENU Forma d'onda arbitraria Nome del soft-key Arb Parametro Selezione onda Nome del soft-key Waves Impostazione Sinc Frequenza - 10kHz Ampiezza Ampl 2V Offset Offset 5mV Parametro Stato uscita NOME DEL TASTO FISSO Output1 Impostazione On 152 TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore Configurazione di una forma d’onda sinusoidale modulata AM Configurazione di una forma d’onda sinusoidale modulata AM Iniziare con lo strumento configurato con le impostazioni predefinite. Parametro Nome del soft-key Frequenza - Impostazione 10MHz MENU NOME DEL TASTO FISSO Modulazione Mod Parametro Frequenza Nome del soft-key - Impostazione 100kHz Profondità Depth 100% Sorgente Source Internal Forma Shape Sine Stato modulazione On/Off On Parametro NOME DEL TASTO FISSO Impostazione Stato uscita Output1 On Cambio della modulazione in AM-SC Parametro Nome del soft-key Tipo Type > AM 153 Impostazione AM-SC TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore PRBS PRBS Iniziare con lo strumento configurato con le impostazioni predefinite. MENU Forma d'onda arbitraria NOME DEL TASTO FISSO Waves MENU Rumore/PRBS Nome del soft-key Noise Impostazione Sorgente Source PBRS Parametro Velocità di trasmissione Nome del soft-key BitRate Impostazione 1Mbps Ampiezza Ampl 3,3V Offset Offset 1,65V Tipo PRBS Type PN7 Parametro NOME DEL TASTO FISSO Impostazione Stato uscita Output1 On 154 TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore Modulazione di frequenza di una forma d’onda sinusoidale Modulazione di frequenza di una forma d’onda sinusoidale Iniziare con lo strumento configurato con le impostazioni predefinite. MENU Modulazione NOME DEL TASTO FISSO Mod Parametro Stato modulazione Nome del soft-key On/Off Impostazione On Tipo modulazione Type FM Frequenza di modulazione - 1kHz Deviazione Deviatn 9kHz MENU Sinusoide NOME DEL TASTO FISSO Params Parametro Ampiezza Nome del soft-key Ampl Impostazione 1,0V Offset Offset 0,0V Parametro NOME DEL TASTO FISSO Impostazione Stato uscita Output1 On 155 TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore Forma d’onda modulata ad ampiezza di impulso (PWM) Forma d’onda modulata ad ampiezza di impulso (PWM) Iniziare con lo strumento configurato con le impostazioni predefinite. MENU Forma d'onda arbitraria NOME DEL TASTO FISSO Waves Parametro Impulso Nome del soft-key Pulse MENU NOME DEL TASTO FISSO Modulazione Mod Parametro Stato modulazione Nome del soft-key On/Off Impostazione On Tipo modulazione Type PWM Frequenza di modulazione - 1kHz Deviazione Dev % 40% MENU Impulso NOME DEL TASTO FISSO Params Parametro Ampiezza Nome del soft-key Ampl Impostazione 1,0V Offset Offset 0,0V Parametro Stato uscita NOME DEL TASTO FISSO Output1 Impostazione On 156 Impostazione TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore Amplitude Shift Keying (ASK) (Modulazione a spostamento di ampiezza) Amplitude Shift Keying (ASK) (Modulazione a spostamento di ampiezza) Iniziare con lo strumento configurato con le impostazioni predefinite. MENU Modulazione NOME DEL TASTO FISSO Mod Parametro Stato modulazione Nome del soft-key On/Off Impostazione On Tipo modulazione Type ASK Sorgente di modulazione Source Internal Ampiezza Hop HpAmpl 100mV Velocità di commutazione Rate 1kHz Polarità Hop HopPol Positive MENU Sinusoide NOME DEL TASTO FISSO Params Parametro Ampiezza Nome del soft-key Ampl Impostazione 1,0V Offset Offset 0,0V Parametro Stato uscita NOME DEL TASTO FISSO Output1 Impostazione On Notare che la seconda traccia è l’uscita dal socket principale dell’uscita 2 quando il canale 2 è sincronizzato, che segue il segnale di forma d’onda di modulazione. 157 TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore Frequenza di scansione di un’onda sinusoidale Frequenza di scansione di un’onda sinusoidale Iniziare con lo strumento configurato con le impostazioni predefinite. MENU Scansione NOME DEL TASTO FISSO Sweep Parametro Stato scansione Nome del soft-key On/Off Impostazione On Frequenza arresto Freq > Stop 100kHz MENU Sinusoide NOME DEL TASTO FISSO Params Parametro Ampiezza Nome del soft-key Ampl Impostazione 1,0V Offset Offset 0,0V Parametro Stato uscita NOME DEL TASTO FISSO Output1 Impostazione On 158 TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore Generazione di un burst triggered Generazione di un burst triggered Iniziare con lo strumento configurato con le impostazioni predefinite. Parametro Frequenza Nome del soft-key - Impostazione 6MHz MENU Burst NOME DEL TASTO FISSO Burst Parametro Stato burst Nome del soft-key On/Off Impostazione On Conteggio dei burst Count 3 Origine trigger SetTrg > Source > Int Internal Trigger Periodo trigger SetTrg > Period 5ms MENU Sinusoide NOME DEL TASTO FISSO Params Parametro Ampiezza Nome del soft-key Ampl Impostazione 1,0V Offset Offset 0,0V Parametro Stato uscita NOME DEL TASTO FISSO Output1 Impostazione On Notare che la seconda traccia è l’uscita dal socket principale dell’uscita 2 quando il canale 2 è sincronizzato, che segue il segnale di forma di ingresso trigger. 159 TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore Accoppiamento della frequenza di entrambi i canali Accoppiamento della frequenza di entrambi i canali Iniziare con lo strumento configurato con le impostazioni predefinite. MENU Utility NOME DEL TASTO FISSO Utility Parametro Frequenze Nome del soft-key Dual Ch > Freq Impostazione Coupled Parametro Stato uscita NOME DEL TASTO FISSO Output1 Impostazione On Stato uscita Output2 On MENU Sinusoide NOME DEL TASTO FISSO Parmas Parametro Spostamento di fase Nome del soft-key Output > Phase > Done Impostazione 90 gradi Frequenza Freq 1MHz Se il canale 1 è impostato su 1MHz, anche il canale 2 è impostato su 1MHz. Si può vedere lo spostamento di fase di 90 gradi tra i canali. 160 TGF4000 Series Quick Start Guide 5 - Analisi delle funzionalità del generatore Contatore di frequenza Contatore di frequenza Iniziare con lo strumento configurato con le impostazioni predefinite. MENU Utility NOME DEL TASTO FISSO Utility Parametro Contatore Nome del soft-key Instr > FrCntr > On/Off Impostazione Counter Enabled Sorgente Source TRIG IN - DC coupled Tipo Freq Frequency Misurazione Count La casella di modifica mostra la misurazione corrente. Se non è presente alcun segnale in entrata all’ingresso selezionato, il contatore segna ’No signal’ finché non viene applicato un segnale in entrata adeguato. Quando viene applicato un segnale di ingresso adeguato, il contatore misura costantemente e visualizza il valore nella casella di modifica. 161 TGF4000 Series Quick Start Guide 6 - Manutenzione Pulizia 6. MANUTENZIONE I Produttori o i loro agenti all’estero faranno le riparazioni necessarie in caso di guasto. The Manufacturers or their agents overseas will provide a repair service for any unit developing a fault. Nel caso in cui i proprietari desiderano effettuare da sé i lavori di manutenzione, questi devono essere svolti esclusivamente da personale qualificato in conformità con il manuale di servizio, che può essere ottenuto direttamente dai produttori o dai propri agenti distaccati. Pulizia Se si deve pulire il dispositivo di alimentazione, usare uno strofinaccio appena bagnato con acqua o con un detergente neutro. Pulire la finestrella di visualizzazione con un panno asciutto e morbido. ATTENZIONE! PER EVITARE SCOSSE ELETTRICHE ED EVENTUALI DANNI AL DISPOSITIVO DI ALIMENTAZIONE, NON PERMETTERE MAI ALL’ACQUA DI ENTRARE ALL’INTERNO DELL’ALLOGGIAMENTO. PER EVITARE DANNI ALL’ALLOGGIAMENTO E ALLA FINESTRELLA DI VISUALIZZAZIONE, NON PULIRE MAI CON SOLVENTI. 162 TGF4000 Series Quick Start Guide 0 - Índice ÍNDICE 1. Seguridad......................................................................................................................................................... 164 Símbolos ............................................................................................................................................................... 165 2. Principios de funcionamiento .......................................................................................................................... 166 Distribución del panel frontal ............................................................................................................................... 166 Distribución del panel posterior ........................................................................................................................... 167 Distribución de la pantalla .................................................................................................................................... 168 3. Primeros pasos ................................................................................................................................................ 169 Estado inicial......................................................................................................................................................... 169 4. Ejemplos básicos de configuración................................................................................................................... 170 Configurar una señal de onda senoidal ................................................................................................................ 170 Configurar una señal de reloj de onda cuadrada.................................................................................................. 174 Configurar una forma de onda de pulso ............................................................................................................... 180 Configurar más opciones de salida ....................................................................................................................... 188 5. Repaso de las capacidades del generador ........................................................................................................ 192 Configurar una señal de onda arbitraria ............................................................................................................... 192 Configurar una forma de onda senoidal con modulación AM .............................................................................. 193 SBSA ..................................................................................................................................................................... 194 Modulación de frecuencia de una forma de onda de senoidal ............................................................................ 194 Forma de onda modulada por anchura de pulso (PWM) ..................................................................................... 196 Modulación por desplazamiento de amplitud (ASK) ............................................................................................ 196 Barrido de frecuencia de onda senoidal ............................................................................................................... 198 Generación de una ráfaga activada ...................................................................................................................... 199 Emparejamiento de la frecuencia de ambos canales ........................................................................................... 200 Contador de frecuencia ........................................................................................................................................ 201 6. Mantenimiento ................................................................................................................................................ 202 Limpieza ............................................................................................................................................................... 202 7. Specification .................................................................................................................................................... 203 Las últimas revisiones de este manual, controladores de dispositivos y herramientas de software se pueden descargar desde: www.aimtti.com 163 TGF4000 Series Quick Start Guide 1 - Seguridad 1. SEGURIDAD El presente generador constituye un instrumento que pertenece a la Clase de Seguridad I de la clasificación CEI y ha sido diseñada para cumplir las prescripciones de la norma EN 61010−1 (Requisitos de seguridad de equipos eléctricos de medida, control y uso en laboratorio). Se trata de un instrumento de la Categoría de Instalación II que se debe alimentar con un suministro monofásico normal. Este instrumento se ha sometido a pruebas con arreglo a la norma EN 61010−1 y se proporciona en condiciones de funcionamiento seguro. El presente manual de instrucciones contiene información y advertencias que el usuario debe seguir, con el fin de garantizar y perpetuar la seguridad de funcionamiento. Este instrumento ha sido diseñado para su uso en interiores, en entornos con una polución de grado 2 y dentro de un intervalo de temperaturas comprendido entre 5 °C y 40 °C, con una humedad relativa comprendida entre el 20 % y el 80 % (sin condensación). Se puede someter ocasionalmente a temperaturas comprendidas entre +5 °C y −10 °C, sin que su seguridad se vea reducida. No se debe utilizar cuando haya condensación. El uso de este instrumento de forma distinta a la especificada en estas instrucciones puede afectar a sus mecanismos de seguridad. No utilice el instrumento con tensiones ni en entornos que se encuentren fuera del intervalo especificado. ¡ADVERTENCIA! ESTE INSTRUMENTO DEBE CONECTARSE A TIERRA Cualquier interrupción del conductor de puesta a tierra, dentro o fuera del instrumento, hará que este resulte peligroso. Está prohibida la interrupción intencionada. No se debe inhibir este mecanismo de protección mediante un alargador que no tenga conductor de toma de tierra. Cuando el instrumento esté conectado a la fuente de alimentación puede haber terminales con tensión y es probable que, si se abre la carcasa o se retiran piezas a las que no sea posible acceder manualmente en condiciones normales, queden al descubierto componentes con tensión. Es necesario desconectar el instrumento de cualquier fuente de alimentación antes de abrirlo para realizar tareas de ajuste, sustitución, mantenimiento o reparación. Se debe evitar en la medida de lo posible la realización de cualquier tarea de ajuste, mantenimiento o reparación del instrumento abierto con tensión y, si fuera inevitable, solo la realizará una persona con la preparación suficiente y que conozca los peligros inherentes.Si el instrumento resultara estar claramente defectuoso, hubiera sido sometido a un daño mecánico, a humedad excesiva o a corrosión química, su protección de seguridad podría fallar, por lo que será necesario dejar de utilizar el aparato y devolverlo para su comprobación y reparación. Asegúrese de que, cuando sea necesaria una sustitución, se utilicen únicamente fusibles de la corriente nominal y el tipo especificados. Está prohibido utilizar fusibles artesanales o cortocircuitar los portafusibles. Este instrumento usa una pila de botón de litio como batería−de respaldo para la memoria no volátil−; su duración típica es de 5 años. De ser necesario, sustitúyala solo por otra del tipo correcto: una pila de botón de Li/MnO2 de 3 V, tipo 2032 (20 mm). Las pilas agotadas deben eliminarse de forma respetuosa y de acuerdo con la legislación local; no las abra, incinere ni exponga a temperaturas por encima de 60 °C, y no intente recargarlas. No se debe humedecer el instrumento durante la limpieza; en particular, la pantalla LCD se debe limpiar únicamente con un paño suave y seco. 164 TGF4000 Series Quick Start Guide 1 - Seguridad Símbolos Símbolos En el instrumento y en este manual se utilizan los siguientes símbolos:− Precaución −Remítase a los documentos adjuntos; un uso incorrecto puede dañar el instrumento. Terminal conectado a tierra. Alimentación de red OFF (desconectada). l Alimentación de red ON (conectada). Corriente alterna (CA). 165 TGF4000 Series Quick Start Guide 2 - Principios de funcionamiento Distribución del panel frontal 2. PRINCIPIOS DE FUNCIONAMIENTO Distribución del panel frontal Ref. Descripción breve Función 1 Interruptor de alimentación Enciende y apaga el instrumento. Nota de seguridad: Para desconectarlo completamente de la corriente alterna, desenchufe el cable de alimentación en la parte posterior del instrumento o en el punto de la toma de red; asegúrese de tener siempre accesible un medio de desconexión. 2 Teclas variables Ejecutan la función indicada en la etiqueta situada en la pantalla LCD sobre ella. 3 Menús de forma de onda Seleccionan el tipo principal de forma de onda activa, es decir, la forma de onda portadora (senoidal, cuadrada, rampa, pulso, ruido/SBSA o arbitraria). 4 Menús de modificación de forma de onda Abre los menús de configuración de los parámetros de modulación, barrido y ráfaga. 5 Otros menús Selecciona los menús de almacenamiento interno y externo de archivos, utilidades del instrumento y condiciones de activación. 6 Tomas principales Tomas principales de salida. El canal 2 también puede configurarse para transportar la sincronía del canal 1 desde su toma MAIN OUT 2. 7 Teclas de salida Activan o desactivan la salida principal MAIN OUT seleccionada. 8 Teclas de cursor y mando giratorio. Usadas para cambiar los valores de los parámetros numéricos dígito a dígito. Usadas para seleccionar elementos en algunos menús. 9 Teclado numérico Usado para introducir valores numéricos directamente. 10 Almacenamiento USB Conector host USB para unidad de memoria USB. 166 TGF4000 Series Quick Start Guide 2 - Principios de funcionamiento Distribución del panel posterior Distribución del panel posterior Ref. Descripción breve Función 1 Entrada de modulación Entrada externa para modulación de las formas de onda principales. 2 Entrada de referencia/Contador de frecuencia con acoplamiento de CA Entrada para reloj de referencia a 10 MHz y medición externa de frecuencia con acoplamiento de CA. 3 Salida de referencia Salida para reloj de referencia interna a 10 MHz. 4 Entrada de activación/Contador de frecuencia con acoplamiento de CC Entrada externa para activar las formas de onda principales y medición externa de frecuencia con acoplamiento de CC. 5 LAN (red de área local) La interfaz LAN está diseñada para ser compatible con el estándar LXI Core 2011. El control remoto a través de la interfaz LAN es posible mediante el protocolo de sockets de TCP/IP. 6 USB El puerto USB, que acepta un cable USB estándar. 7 GPIB (IEEE−488): opcional La dirección GPIB por defecto es 5. 167 TGF4000 Series Quick Start Guide 2 - Principios de funcionamiento Distribución de la pantalla Distribución de la pantalla Ref. Descripción breve Función 1 Indicador de canal Muestra el canal actualmente seleccionado 2 Tipo de forma de onda principal Muestra la forma de onda portadora actual 3 Estado de salida Muestra si la salida principal está encendida o apagada. 4 Indicador de señal de reloj externa Muestra el estado de la señal de reloj externa (si la hubiera). 5 Indicador del estado de la LAN Muestra el estado de la conexión de área local (Ethernet). 6 Caja de parámetros Muestra los principales parámetros de la forma de onda. 7 Descripción del menú Muestra el menú de edición actualmente seleccionado. 8 Caja de gráficas Muestra una representación gráfica de la forma de onda seleccionada. 9 Caja de edición Muestra el parámetro actual que puede ser editado. 10 Etiquetas de teclas variables Muestra las funciones actuales de las seis teclas situadas debajo. 168 TGF4000 Series Quick Start Guide 3 - Primeros pasos Estado inicial 3. PRIMEROS PASOS A fin de que el usuario se familiarice con algunas de las funcionalidades básicas del instrumento, esta guía presenta una serie de ejemplos de configuración. Se recomienda a todos los usuarios proceder con los tres primeros ejemplos: • Configurar una señal de onda senoidal • Configurar una señal de reloj de onda cuadrada • Configurar una forma de onda de pulso • Configurar más opciones de salida Se incluyen una serie de ejemplos de configuración adicionales que asumen cierta familiarización con el instrumento. • Configurar una señal de onda arbitraria • Configurar una forma de onda senoidal con modulación AM • SBSA (TGF4162 & TGF4242) • Modulación de frecuencia de una forma de onda de senoidal • Forma de onda modulada por anchura de pulso (PWM) • Modulación por desplazamiento de amplitud (ASK) • Barrido de frecuencia de onda senoidal • Generación de una ráfaga activada • Emparejamiento de la frecuencia de ambos canales • Contador de frecuencia Estado inicial Antes de configurar el instrumento para cualquiera de los ejemplos, deberá restablecerse su estado por defecto. Para ello, lleve a cabo los siguientes pasos: • • • • Pulse la tecla física Utility (utilidad). Pulse la tecla variable System (sistema). Pulse la tecla variable Default (por defecto): la pantalla mostrará Restore Factory Default? (¿restaurar valores de fábrica por defecto?). Pulse la tecla variable Yes (sí). Estas acciones establecen el seno como forma de onda principal (10 kHz, 1 Vpp) y cancelan cualquier modulación, barrido o activación/conmutación de ráfaga. NOTA: Es posible configurar el instrumento para que recuerde sus últimos valores al apagarlo y los restaure al encenderse. Esto se realiza desde el menú Utility, con la tecla variable PwrOn (encendido). Esta configuración se perderá cuando se restablezca el estado por defecto del instrumento según lo anteriormente descrito. 169 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de onda senoidal 4. EJEMPLOS BÁSICOS DE CONFIGURACIÓN Configurar una señal de onda senoidal Requisito Emitir una señal de onda senoidal con una frecuencia de 40 MHz y amplitud de 6 Vpp en la salida MAIN OUT 1. Estado inicial Antes de empezar, restaure los valores por defecto del instrumento según lo descrito en la secciónEstado inicial Abrir menú de forma de onda senoidal • Pulse la tecla física Waves • Pulse la tecla variable Sine 170 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de onda senoidal Configurar la frecuencia Se resaltará la tecla variable Freq: en la caja de edición se mostrará la frecuencia actual. Tenga en cuenta que, al pulsar repetidamente esta tecla variable, su función alterna entre Frequency y Period. • Use el teclado numérico para introducir una nueva frecuencia. Pulse los números 4 0 Advierta cómo, tan pronto como los introduce, las teclas variables pasan a mostrar las unidades de frecuencia. • 171 Pulse la tecla variable MHz para confirmar una frecuencia de 40 MHz. TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de onda senoidal Configurar la amplitud • Pulse la tecla variable Ampl. Advierta cómo las sucesivas pulsaciones de la tecla variable Ampl alternan las etiquetas Ampl y Offset (desplazamiento) con HiLvl (nivel alto) y LoLvl (nivel bajo) y viceversa. • Use el teclado numérico para introducir una nueva amplitud. Pulse el número 6 Advierta cómo, tan pronto como lo introduce, las teclas variables pasan a mostrar las unidades de tensión. 172 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de onda senoidal • Pulse la tecla variable Vpp para confirmar una amplitud pico a pico de 6,0 voltios. Encender la salida • Pulse la tecla Output 1 para encender la salida del canal 1. La tecla de salida 1 se ilumina en naranja para indicar el estado de encendido 173 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de reloj de onda cuadrada Configurar una señal de reloj de onda cuadrada Requisito Emitir una señal de reloj de onda cuadrada continua con una frecuencia de 20 MHz, ciclo de trabajo del 50 %, nivel alto de 3,3 V y nivel bajo de 0,0 V desde MAIN OUT 1. Estado inicial Antes de empezar, restaure los valores por defecto del instrumento según lo descrito en la sección Estado inicial Abrir menú de forma de onda cuadrada • Pulse la tecla física Waves • Pulse la tecla variable Square 174 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de reloj de onda cuadrada Configurar la frecuencia Se resaltará la tecla variable Freq: en la caja de edición se mostrará la frecuencia actual. Tenga en cuenta que, al pulsar repetidamente esta tecla variable, su función alterna entre Frequency y Period. • Use el teclado numérico para introducir una nueva frecuencia. Pulse los números 2 0 Advierta cómo, tan pronto como los introduce, las teclas variables pasan a mostrar las unidades de frecuencia. • Pulse la tecla variable MHz para confirmar una frecuencia de 20 MHz. Advierta cómo la caja de gráficas cambia para mostrar el tiempo de ascenso en los flancos, por ser ahora un aspecto significativo. 175 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de reloj de onda cuadrada Confirmar el ciclo de trabajo • Pulse la tecla variable Duty (ciclo de trabajo): en la caja de edición se mostrará el ciclo de trabajo actual. Observe que el ciclo de trabajo ya está configurado al 50 %, pero puede cambiarse desde aquí si lo desea. Configurar el nivel alto y bajo • Pulse la tecla variable Ampl; su etiqueta cambia a HiLvl y en la caja de edición se muestra la tensión actual del nivel alto. Advierta cómo las sucesivas pulsaciones de la tecla variable Ampl alternan las etiquetas Ampl y Offset (desplazamiento) con HiLvl (nivel alto) y LoLvl (nivel bajo) y viceversa. 176 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de reloj de onda cuadrada Cuando la tecla variable indica HiLvl, en la caja de edición se muestra la tensión actual del nivel alto. • Use el teclado numérico para introducir un nivel distinto. Pulse los números 3 . 3 Advierta cómo, tan pronto como lo introduce, las teclas variables pasan a mostrar las unidades de tensión. • 177 Pulse la tecla variable V para confirmar un nivel alto de 3,3 voltios. TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de reloj de onda cuadrada • Pulse la tecla variable LoLvl - en la caja de edición se muestra la tensión actual del nivel bajo. • Use el teclado numérico para introducir un nivel distinto. Pulse 0 • Pulse la tecla variable V para confirmar un nivel bajo de 0,0 voltios. Encender la salida • Pulse la tecla Output1 para encender la salida del canal 1. La tecla de salida 1 se ilumina en naranja para indicar el estado de encendido. 178 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una señal de reloj de onda cuadrada Cambio de cualquier parámetro numérico en tiempo real (p. ej. la frecuencia) Los parámetros numéricos pueden cambiarse usando las teclas de cursor y el mando giratorio como alternativa al teclado numérico. • Pulse la tecla física Waves • Pulse la tecla variable Square • Pulse la tecla variable Freq: se mostrará la frecuencia actual de 20,0 MHz. • Pulse las teclas físicas del cursor para mover el foco de edición al segundo dígito. • Use el mando giratorio para modificar el valor: la frecuencia se cambia de inmediato. NOTA: Presione para activar / desactivar la rueda de giro. 179 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una forma de onda de pulso Configurar una forma de onda de pulso Requisito Emitir una señal de pulso continua con un periodo de 100 ns, anchura de pulso de 30 ns, tiempos de flanco de 20 ns, nivel alto de 2,7 V y nivel bajo de -0,6 V desde MAIN OUT 1. Estado inicial Antes de empezar, restaure los valores por defecto del instrumento según lo descrito en la sección Estado inicial Abrir menú de forma de onda de pulso • Pulse la tecla física Waves • Pulse la tecla variable Pulse. 180 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una forma de onda de pulso Configurar el periodo • Pulse la tecla variable PlsFreq (frecuencia de pulso) de modo que cambie a PlsPer (periodo de pulso): la caja de edición pasará a mostrar el periodo actual. Tenga en cuenta que, al pulsar repetidamente esta tecla, su función alterna entre Frequency y Period. • Use el teclado numérico para introducir un periodo distinto. Pulse los números 1 0 0 Advierta cómo, tan pronto como los introduce, las teclas variables pasan a mostrar las unidades de tiempo. 181 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una forma de onda de pulso • Pulse la tecla variable ns para confirmar un periodo de 100 ns. Observe cómo la caja de gráficas pasa a mostrar una representación de los tiempos de pulso y flanco. Configurar la anchura del pulso • 182 Pulse la tecla variable Duty (ciclo de trabajo); su etiqueta cambia a Width (anchura) y se muestra la anchura en unidades de tiempo. TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una forma de onda de pulso • Use el teclado numérico para introducir una nueva anchura. Pulse los números 3 0 Advierta cómo, tan pronto como los introduce, las teclas variables pasan a mostrar las unidades de tiempo. • 183 Pulse la tecla variable ns para confirmar una anchura de 30 ns. TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una forma de onda de pulso Configurar los tiempos de flanco del pulso • Pulse la tecla variable More • Pulse la tecla variable Edge • Use las teclas de cursor para seleccionar el valor que representa las unidades de 10 ns. • Use el mando giratorio para cambiar el valor a 20,0 ns. • Pulse la tecla variable Done. Observe cómo el valor podría haberse introducido utilizando el teclado numérico si así lo desea. 184 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una forma de onda de pulso Configurar el nivel alto y bajo • Pulse la tecla variable Ampl; su etiqueta cambia a HiLvl y en la caja de edición se muestra la tensión actual del nivel alto. Advierta cómo las sucesivas pulsaciones de la tecla variable Ampl alternan las etiquetas Ampl y Offset (desplazamiento) con HiLvl (nivel alto) y LoLvl (nivel bajo) y viceversa. • Use el teclado numérico para introducir un nivel distinto. Pulse los números 2 . 7 . Advierta cómo, tan pronto como lo introduce, las teclas variables pasan a mostrar las unidades de tensión. 185 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una forma de onda de pulso • Pulse la tecla variable V para confirmar un nivel alto de 2,7 voltios. • Pulse la tecla variable LoLvl: en la caja de edición se muestra la tensión actual del nivel bajo. 186 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar una forma de onda de pulso • Use el teclado numérico para introducir un nivel distinto. Pulse - . 6 • Pulse la tecla variable V para confirmar un nivel bajo de -600 mV. Encender la salida • Pulse la tecla Output1 para encender la salida del canal 1. La tecla de salida 1 se ilumina en naranja para indicar el estado de encendido 187 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar más opciones de salida Configurar más opciones de salida Requisito En los ejemplos de configuración previos se ha mostrado cómo usar el menú de salida para configurar el nivel de la salida (amplitud más desplazamiento, o nivel alto más nivel bajo) y encenderla o apagarla. Este ejemplo demuestra la configuración de la fase de salida, la polaridad de salida, la impedancia de la carga y el intervalo automático de la tensión. Estado inicial Antes de empezar, restaure los valores por defecto del instrumento según lo descrito en la sección Estado inicial Abrir el menú de salida • 188 Pulse la tecla variable Output. TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar más opciones de salida Cambiar la fase de salida Se seleccionará por defecto la tecla variable Phase. • Introduzca una fase de - 4 5 ° El ángulo de fase configurado es el punto del periodo de la forma de onda que coincide con el flanco de activación o sincronismo; es decir, el punto del periodo en el que comienza la forma de onda. Consecuentemente, al configurarse una fase negativa, la forma de onda se adelanta, mientras que con una fase negativa se retrasa, con respecto a la activación o el sincronismo; la forma de onda de la caja de gráficas se actualiza para mostrar este cambio. La tecla variable Align (alinear) se emplea para realinear la fase al realizar cambios en la frecuencia. Cambiar la polaridad de salida • Pulse la tecla variable Type (tipo) para invertir la polaridad de la salida. Tenga en cuenta que, pulsando sucesivamente la tecla Type, se alterna entre polaridad normal e invertida. 189 TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar más opciones de salida Cambiar la impedancia de carga • Pulse la tecla variable Load (carga). • Vuelva a pulsar la tecla variable Load para modificar la impedancia a High-z (alta impedancia). Advierta que, aunque el valor por defecto de la impedancia de carga es 50 ohmios, puede cambiarse a cualquier otro entre 50 y 10.000 ohmios. Los niveles se calculan en función de esta impedancia. Al pulsar sucesivamente la tecla Load, su función alterna entre un valor numérico y High-z. Observe cómo la lectura de la amplitud aumenta hasta 2 voltios pico a pico. • 190 Pulse la tecla variable Load para devolver la impedancia de la carga a 50 ohmios. TGF4000 Series Quick Start Guide 4 - Ejemplos básicos de configuración Configurar más opciones de salida Cambiar el intervalo • Pulse la tecla variable Range (intervalo). • Vuelva a pulsar la tecla variable Range para modificar el intervalo de Auto (automático) a Hold (fijo). El modo Auto varía automáticamente el intervalo en pasos de 6 dB de atenuación (es decir, máximos de intervalo de 10 Vpp, 5 Vpp, 2,5 Vpp, etc., en 50 Ω), con el intervalo de calibrado de la amplitud limitado a 6 dB para mantener la calidad de la forma de onda. Al seleccionar el modo Hold se deshabilita el intervalo automático; el valor de atenuación pasa a ser fijo y el intervalo de calibrado de la amplitud deja de estar limitado. Con el intervalo configurado en Auto, la amplitud y los atenuadores cambian automáticamente para lograr un rendimiento óptimo. Con el intervalo configurado en Hold, se utiliza un valor de atenuación fijo para todos los valores de amplitud. 191 TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador Configurar una señal de onda arbitraria 5. REPASO DE LAS CAPACIDADES DEL GENERADOR En los siguientes ejemplos se indica únicamente la configuración de los parámetros, así como los nombres de las teclas relacionadas. Se presentan las formas de onda de salida resultantes junto a la forma de onda de sincronismo o activación cuando proceda. La amplitud de la salida y la configuración del desplazamiento son tan solo ejemplos que no es necesario observar. Configurar una señal de onda arbitraria Comience con el instrumento de nuevo en la configuración por defecto. MENÚ Tecla variable Forma de onda arbitraria Waves MENÚ Forma de onda arbitraria TECLA FÍSICA Arb Parámetro Selección de onda Tecla variable Waves Configuración Sinc Frecuencia - 10kHz Amplitud Ampl 2V Desviación Offset 5mV Parámetro Estado de salida TECLA FÍSICA Output 1 Configuración On 192 TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador Configurar una forma de onda senoidal con modulación AM Configurar una forma de onda senoidal con modulación AM Comience con el instrumento de nuevo en la configuración por defecto. Parámetro Tecla variable Configuración Frecuencia 10MHz MENÚ Modulación TECLA FÍSICA Mod Parámetro Frecuencia Tecla variable - Configuración 100kHz Profundidad Depth 100% Fuente Source Internal Forma Shape Sine Estado de modulación On/Off On Parámetro Estado de salida TECLA FÍSICA Output 1 Configuración On Tecla variable Type > AM Configuración AM-SC Cambiar la modulación a AM-SC Parámetro Tipo 193 TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador SBSA SBSA Comience con el instrumento de nuevo en la configuración por defecto. MENÚ Forma de onda arbitraria TECLA FÍSICA Waves MENÚ Ruido/SBSA Tecla variable Noise Configuración Fuente Source PRBS Parámetro Tasa de bits Tecla variable BitRate Configuración 1Mbps Amplitud Ampl 3.3V Desviación Offset 1.65V Tipo de SBSA Type PN7 Parámetro Estado de salida TECLA FÍSICA Output 1 Configuración On 194 TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador Modulación de frecuencia de una forma de onda de senoidal Modulación de frecuencia de una forma de onda de senoidal Comience con el instrumento de nuevo en la configuración por defecto. MENÚ Modulación TECLA FÍSICA Mod Parámetro Estado de modulación Tecla variable On/Off Configuración On Tipo de modulación Type FM Frecuencia de modulación - 1kHz Desviación Deviatn 9kHz MENÚ Senoidal TECLA FÍSICA Params Parámetro Amplitud Tecla variable Ampl Configuración 1.0V Desviación Offset 0.0V Parámetro Estado de salida TECLA FÍSICA Output 1 Configuración On 195 TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador Forma de onda modulada por anchura de pulso (PWM) Forma de onda modulada por anchura de pulso (PWM) Comience con el instrumento de nuevo en la configuración por defecto. MENÚ Forma de onda arbitraria TECLA FÍSICA Waves Parámetro Pulso Tecla variable Pulse MENÚ TECLA FÍSICA Modulación Mod Parámetro Estado de modulación Tecla variable On/Off Configuración On Tipo de modulación Type PWM Frecuencia de modulación - 1kHz Desviación Dev % 40% MENÚ Pulso TECLA FÍSICA Params Parámetro Amplitud Tecla variable Ampl Configuración 1.0V Desviación Offset 0.0V Parámetro Estado de salida TECLA FÍSICA Output 1 Configuración On 196 Configuración TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador Modulación por desplazamiento de amplitud (ASK) Modulación por desplazamiento de amplitud (ASK) Comience con el instrumento de nuevo en la configuración por defecto. MENÚ Modulación TECLA FÍSICA Mod Parámetro Estado de modulación Tecla variable On/Off Configuración On Tipo de modulación Type ASK Fuente de modulación Source Internal Amplitud de salto HpAmpl 100mV Tasa de conmutación Rate 1kHz Polaridad del salto HopPol Positive MENÚ Senoidal TECLA FÍSICA Params Parámetro Amplitud Tecla variable Ampl Configuración 1.0V Desviación Offset 0.0V Parámetro Estado de salida TECLA FÍSICA Output 1 Configuración On Advierta que la segunda traza representa la salida de la toma Main Output 2 cuando el canal 2 está sincronizado: dicha salida sigue la señal de la onda moduladora. 197 TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador Barrido de frecuencia de onda senoidal Barrido de frecuencia de onda senoidal Comience con el instrumento de nuevo en la configuración por defecto. MENÚ Barrido TECLA FÍSICA Sweep Parámetro Estado de barrido Tecla variable On/Off Configuración On Frecuencia de paro Freq > Stop 100kHz MENÚ Senoidal TECLA FÍSICA Params Parámetro Amplitud Tecla variable Ampl Configuración 1.0V Desviación Offset 0.0V Parámetro Estado de salida TECLA FÍSICA Output 1 Configuración On 198 TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador Generación de una ráfaga activada Generación de una ráfaga activada Comience con el instrumento de nuevo en la configuración por defecto. Parámetro Frecuencia Tecla variable - Configuración 6MHz MENÚ Ráfaga TECLA FÍSICA Burst Parámetro Estado de ráfaga Tecla variable On/Off Configuración On Recuento de ráfaga Count 3 Fuente de activación SetTrg > Source > Int Internal Trigger Periodo de activación SetTrg > Period 5ms MENÚ Senoidal TECLA FÍSICA Params Parámetro Amplitud Tecla variable Ampl Configuración 1.0V Desviación Offset 0.0V Parámetro Estado de salida TECLA FÍSICA Output 1 Configuración On Advierta que la segunda traza representa la salida de la toma Main Output 2 cuando el canal 2 está sincronizado: dicha salida sigue la señal de entrada de activación. 199 TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador Emparejamiento de la frecuencia de ambos canales Emparejamiento de la frecuencia de ambos canales Comience con el instrumento de nuevo en la configuración por defecto. MENÚ Utilidad TECLA FÍSICA Utility Parámetro Frecuencias Tecla variable Dual Ch >Freq Configuración Coupled Parámetro Estado de salida TECLA FÍSICA Output1 Configuración On Estado de salida Output2 On MENÚ Senoidal TECLA FÍSICA Params Parámetro Desplazamiento de fase Tecla variable Output > Phase > Done Configuración 90° Frecuencia Freq 1MHz Observe cómo, al configurar el canal 1 a 1MHz, el canal 2 también se configura a 1mHz. Puede verse el desplazamiento de fase de 90 grados entre canales. 200 TGF4000 Series Quick Start Guide 5 - Repaso de las capacidades del generador Contador de frecuencia Contador de frecuencia Comience con el instrumento de nuevo en la configuración por defecto. MENÚ Utilidad TECLA FÍSICA Utility Parámetro Contador Tecla variable Instr > FrCntr > On/Off Configuración Counter enabled Fuente Source TRIG IN - DC coupled Tipo Freq Frequency Medición Count La caja de edición muestra la medición actual. Cuando no existe ninguna señal presente en la entrada seleccionada, el contador indica No signal (sin señal) hasta que se aplique la señal adecuada. Cuando se aplica la señal adecuada, el contador mide y muestra de forma continua su lectura en la caja de edición. 201 TGF4000 Series Quick Start Guide 6 - Mantenimiento Limpieza 6. MANTENIMIENTO Los fabricantes o sus agentes en el extranjero ofrecen un servicio de reparación para toda unidad que desarrolle un defecto. Si los propietarios desearan establecer su propio servicio, esto sólo debe realizarse por personas cualificadas en conjunto con el manual de servicio que puede adquirirse directamente del Fabricante o de sus agentes en el extranjero. Limpieza Si la unidad de suministro de fuerza necesita ser limpiada, utilizar un paño brevemente humedecido en agua o en un detergente suave. La ventana de visualización debe lustrarse con un paño suave y seco. 202 TGF4000 Series Quick Start Guide 7 - Specification 7. SPECIFICATION MODEL: TGF4042 TGF4082 TGF4162 TGF4242 Waveforms Sine, Square, Ramp (Variable Symmetry), Triangle (50% Ramp symmetry), Positive Ramp (100% Ramp symmetry), Negative Ramp (0% Ramp symmetry), Pulse, Noise (Gaussian), DC, Sin(x)/x, Exponential Rise, Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Cardiac, Gaussian, Lorentz, DLorentz and 4 User Defined Arbitrary Waveforms. Dozens of useful pre-built arbitrary waveforms are also supplied on the website: www.aimtti.com PRBS Standard waveforms: Sine Frequency range: Frequency resolution: 1µHz to 40MHz 1µHz, 14 digits 10mVp-p to 10Vp ≤50MHz ≤80MHz ≤120MHz ≤240MHz ≤10MHz Amplitude flatness ≤100MHz (1Vp-p relative to ≤160MHz 10 kHz): ≤240MHz ≤10MHz ≤50MHz Harmonic distortion ≤80MHz (1Vp-p) ≤130MHz ≤240MHz Total harmonic distortion DC to 20kHz (typical): Non-harmonic spurii: Phase noise (10MHz, 1Vp-p, 10kHz offset): 10mVp-p to 10Vp-p 10mVp-p to 5Vp-p Output level (into 50Ω): Square Frequency range: Frequency resolution: Output level (into 50Ω): Duty cycle: Rise and fall times (typical): Aberrations (typical): Jitter (RMS): 1µHz to 80MHz -65dBc -113dBc/Hz 10mVp-p to 10Vp-p 10ns, fixed 1µHz to 100MHz 1µHz, 15 digits 10mVp-p to 10Vp-p 10mVp-p to 4Vp-p 3ns, fixed 5ns, fixed ±5% of amplitude <30ps (cycle to cycle) 1µHz to 5MHz 1µHz, 13 digits 10mVp-p to 10Vp-p <0.1% to 100kHz 0.00% to 100.00%, 0.01% resolution Frequency range: Frequency Resolution: 1mHz to 25MHz 1mHz, 11 digits Pulse 203 10mVp-p to 10Vp-p 10mVp-p to 5Vp-p 10mVp-p to 2.5Vp-p 0.05% Frequency range: Frequency resolution: Output level (into 50Ω): Linearity error: Variable symmetry: Aberrations (Typical): 10mVp-p to 10Vp-p 10mVp-p to 5Vp-p 10mVp-p to 2.5Vp-p 0.001% to 99.999%, 0.001% resolution ≤ 4p-p ≥ 4p-p Ramp Output level (into 50Ω): 1µHz to 240MHz ±0.1dB ±0.2dB ±0.6dB ±1.0dB -65dBc -50dBc -40dBc -35dBc -28dBc 1µHz to 25MHz 1µHz, 14 digits ≤50MHz ≤100MHz 1µHz to 160MHz 1µHz, 15 digits ≤ 50MHz ≤ 100MHz 10mVp-p to 10Vp−p ±5% of amplitude (for transition time 10ns) ±3% of amplitude (for transition time 20ns) <±2% of amplitude (for transition time > 40ns) <0.1% to 200kHz 1mHz to 100MHz 1mHz, 12 digits 10mVp-p to 10Vp−p 10mVp-p to 4Vp-p ±5% of amplitude (for transition time 5ns) ±3% of amplitude (for transition time 10ns) <±2% of amplitude (for transition time > 20ns) TGF4000 Series Quick Start Guide 7 - Specification MODEL: Jitter RMS: Rise and Fall Times: Width: Duty: Delay: Arbitrary TGF4042 TGF4082 TGF4162 TGF4242 <30ps (cycle to cycle) Range: ≤ 4Vp-p 3ns to 799.999999989s (10% to 90%) 8ns to 799.999999984s (10% to 90%) Range: > 4Vp-p 5ns to 799.999999989s (10% to 90%) Resolution: 100ps Accuracy: ±500ps ±0.01% of period Rise and Fall times can be independently varied or can be varied together simultaneously. Range: ≤ 4Vp-p 5ns to 999.999999995s 20ns to 999.999999980s Range: > 4Vp-p 10ns to 999.999999990s Resolution: 100ps Accuracy: ±200ps ±0.01% of period 0.001% to 99.999%, 0.01% resolution Range: 0ns to 999.99999996s 0ns to 999.99999998s Resolution: 100ps Accuracy: ±200ps ±0.01% of period Delay can be entered as absolute delay or phase. In built arbitrary waveforms (Sin(x)/x, Exponential Rise, Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Cardiac, Gaussian, Lorentz and DLorentz). Up to 4 user-defined waveforms may be stored in non-volatile memory. Waveforms can be defined by downloading of waveform data via remote interfaces or from the instrument’s front panel. Waveform Memory Size: 8192 points Vertical Resolution: 14 bits 16 bits In built 1µHz to 2MHz 1µHz to 4MHz Frequency Range: User defined 1µHz to 40MHz 1µHz to 80MHz In built 1µHz, 13 digits Frequency Resolution: User defined 1µHz, 14 digits Output Level (into 50Ω): 10mVp-p to 10Vp-p Sampling rate: 400Msa/s 800MSa/s Point to Point Jitter (Typical): 2.5ns 1.25ns <5ns for 100MHz filter Rise and Fall Times: <8ns <8ns for 62.5MHz filter Effective Analogue Bandwidth (-3dB): 50MHz 62.5MHz, 100MHz, User Selectable Noise Gaussian White Noise: Noise can also be used as modulating waveform. Bandwidth (-3dB): 50MHz Noise crest factor (Vp/Vrms): 6.4 Output Level (into 50Ω): 10mVp-p to 10Vpp 100MHz 5.16 Bit Rate Sequence Length: Rise and Fall Times (Typical): Output Level: 1µbps to 100Mbps, 1µbps resolution 2m – 1, where m = 7, 9, 11, 15, 20, 23, 29 or 31 3ns, Fixed 5ns, Fixed 10mVp-p to 10Vpp into 50Ω PRBS (TGF4162 & TGF4242 only) ≤ 4Vp-p > 4Vp-p - Harmonic Output (TGF4162 & TGF4242 only) Harmonic waveforms can be defined and stored in user-defined arbitrary waveform locations. Frequency Range: Frequency Resolution: Harmonic Order: - Harmonic Amplitude: - Harmonic Phase: Output level (into 50Ω): - 1µHz to 80MHz 1µHz, 14 digits 1 to 50, Up to 16 different harmonics order can be defined 0.0% to 100.0% of output amplitude, 0.1% resolution -360.0 to +360.0 degrees, 0.1 degree resolution 10mVp-p to 10Vp-p Internal frequency reference Internal Modulating Frequency: Amplitude Depth: Internal Modulating Frequency: 204 <± 1ppm <± 1ppm first year <1ppm over the specified temperature range TGF4000 Series Quick Start Guide 7 - Specification MODEL: TGF4042 TGF4082 TGF4162 TGF4242 Modulation AM (Amplitude Modulation) Normal & Suppressed Carrier Carrier Waveforms: Maximum Carrier Frequency: Modulation Source: Internal Modulating Waveforms: Internal Modulating Frequency: Amplitude Depth: FM (frequency modulation) Carrier Waveforms: Modulation Source: Internal Modulating Waveforms: Internal Modulating Frequency: Frequency Deviation: PM (phase modulation) Carrier Waveforms: Modulation Source: Internal Modulating Waveforms: Internal Modulating Frequency: Phase Deviation: ASK (Amplitude Shift Keying) Sine, Square, Ramp, Pulse, Noise, Arb PRBS 25MHz, subject to carrier waveform 50MHz, subject to carrier waveform. Internal/External Sine, Square, Positive Ramp, Negative Ramp, Triangle, Gaussian Noise, DC, Sinc, Exponential Rise, Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Gaussian, Lorentz, D-Lorentz, Cardiac and User Defined Arbs PRBS-PN7, PN9, PN11, PN15, PN20, PN23, PN29, PN31 1µHz to 10MHz, 1µHz resolution 0.00% to 100.00%, 0.01% resolution Sine, Square, Ramp, Arb Internal/External Sine, Square, Positive Ramp, Negative Ramp, Triangle, Gaussian Noise, DC, Sinc, Exponential Rise, Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Gaussian, Lorentz, D-Lorentz, Cardiac and User Defined Arbs. PRBS-PN7, PN9, PN11, PN15, PN20, PN23, PN29, PN31 1µHz to 10MHz, 1µHz resolution DC to Fmax/2, 1µHz resolution Sine, Square, Ramp, Arb Internal/External Sine, Square, Positive Ramp, Negative Ramp, Triangle, Gaussian Noise, DC, Sinc, Exponential Rise, Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Gaussian, Lorentz, D-Lorentz, Cardiac and User Defined Arbs. PRBS-PN7, PN9, PN11, PN15, PN20, PN23, PN29, PN31 1µHz to 10MHz, 1µHz resolution -360.000 to +360.000 degrees, 0.001 degree resolution Maximum Carrier Frequency: Source: Internal Modulation: Sine, Square, Ramp, Pulse, Noise, Arb 25MHz, subject to carrier waveform. Internal/External (via TRIG IN) 2mHz to 10MHz (50% duty cycle square) Carrier Waveforms: Source: Internal Modulation: Sine, Square, Ramp, Arb Internal/External (via TRIG IN) 2mHz to 10MHz (50% duty cycle square) Carrier Waveforms: Modulation Source: Internal Modulation: Sine, Square, Ramp, Arb Internal/External (via TRIG IN) 2mHz to 10MHz (50% duty cycle square) Carrier Waveforms: Modulation Source: Pulse Internal/External Sine, Square, Positive Ramp, Negative Ramp, Triangle, Gaussian Noise, DC, Sinc, Exponential Rise, Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Gaussian, Lorentz, D-Lorentz, Cardiac, and User Defined Arbs PRBS-PN7, PN9, PN11, PN15, PN20, PN23, PN29, PN31 1µHz to 10MHz, 1µHz resolution 0% to 100% of pulse width, 0.01% resolution Carrier Waveforms: FSK (Frequency Shift Keying) BPSK (Binary Phase Shift Keying) PWM (Pulse Width Modulation) Internal Modulating Waveforms: Internal Modulating Frequency: Pulse Width Deviation: 205 PRBS 50MHz, subject to carrier waveform. TGF4000 Series Quick Start Guide 7 - Specification MODEL: TGF4042 TGF4082 SUM (Additive Modulation) (TGF4162 & TGF4242 only) Carrier Waveforms: Maximum Carrier Frequency: Modulation Source: - Internal Modulating Waveforms: - Internal Modulating Frequency: Ratio: - TGF4162 TGF4242 Sine, Square, Ramp, Pulse, Noise, PRBS, Arb 50MHz, subject to carrier waveform. Internal/External Sine, Square, Positive Ramp, Negative Ramp, Triangle, Gaussian Noise, DC, Sinc, Exponential Rise, Exponential Fall, Logarithmic Rise, Logarithmic Fall, Haversine, Gaussian, Lorentz, DLorentz, Cardiac, PRBS-PN7, PN9, PN11, PN15, PN20, PN23, PN29, PN31 and User Defined Arbs. 1µHz to 10MHz, 1µHz resolution 0% to 100%, 0.01% resolution Triggered Burst Each active edge of the trigger signal will produce one burst of the waveform. Sine, Square, Ramp, Pulse, Arb: A fixed number of cycles, specified as number of cycles are generated at every trigger event. Noise: Noise is reset to its start condition at every trigger event. Allows generating same random noise sequence. Carrier Waveforms: PRBS: A fixed number of bits, specified as number of cycles are generated at every trigger event 25MHz (finite cycles), Fmax(infinite), subject to 50MHz (finite cycles), Fmax(infinite), subject to Maximum Carrier Frequency: carrier waveform. carrier waveform. Number of Cycles: 1 to 2147483647 and infinite. Internal 2mHz to 25MHz 2mHz to 50MHz Trigger Repetition Rate: External DC to 1MHz Internal from keyboard or trigger generator. Trigger Signal Source: External from TRIG IN or remote interface. -360.000 to +360.000 degrees, 0.001 degree resolution. Phase off-set cannot be set for Noise and Trigger Start/Stop Phase: PRBS waveforms. Gated Waveform will run while the Gate signal is true and stop while false. Sine, Square, Ramp, Pulse, Noise, Arb Carrier Waveforms: PRBS Maximum Carrier Frequency: 25MHz, subject to carrier waveform. 50MHz, subject to carrier waveform. Internal 2mHz to 25MHz 2mHz to 50MHz Trigger Repetition Rate: External DC to 1MHz Internal from keyboard or trigger generator. Gate Signal Source: External from TRIG IN or remote interface. -360.000 to +360.000 degrees, 0.001 degree resolution. Phase offset cannot be set for Noise and Gate Start/Stop Phase: PRBS waveforms Sweep Frequency sweep capability is provided for both standard and arbitrary waveforms Carrier Waveforms: Sine, Square, Ramp, Arb Sweep Mode: Linear or logarithmic, triggered or continuous. Sweep Direction: Up or Down Sweep Range: From 1µHz to Fmax. Phase continuous. Independent setting of the start and stop frequency. Sweep Time: 1µs to 500s (9 digit resolution). The sweep may be free run or triggered from the following sources: Internal from keyboard or trigger Sweep Trigger Source: generator. Externally from TRIG IN input or remote interface. External trigger repetition rate: DC to 1MHz 206 TGF4000 Series Quick Start Guide 7 - Specification MODEL: TGF4042 TGF4082 TGF4162 TGF4242 Trigger Generator 20ns to 500s or 2mHz to 50MHz square waves 40ns to 500s or 2mHz to 25MHz square waves adjustable in 10ns steps. adjustable in 10ns steps. Resolution: 10ns, 11 digits Each channel has its own trigger generator. Channel 1 trigger is available for external use from the MAIN OUT 2 socket when Channel 2 is configured to output Channel 1 sync waveform and sync source is set to trigger. Internal Source Rate or Frequency: Dual-channel operations Tracking Independent (Off): Equal: The channels are independent of each other. The two channels are identical and behave identically. Coupling Frequency coupling: Amplitude (and DC Offset) coupling: Output coupling: Characteristics Relative phase: Channel to channel Skew (typical): Crosstalk (typical): The frequencies of the two channels can be coupled. Changing the frequency of one channel changes the frequencies of both channels. Amplitude (and DC offset) of the two channels can be coupled. Changing the amplitude and offset on one channel changes the amplitude and offset of both channels. Output On/Off can be coupled. Switching the output On/Off on one channel switches the output On/Off of both channels. -360.000 to 360.000 degrees, 0.001 degree resolution (Phase offset cannot be set for Noise) <1ns (when performing identical operations) <-80db External Frequency Measurement Function: Frequency Range: Source: AC coupled DC coupled AC coupled DC coupled Frequency Resolution: Measurement Time: Input Range and Sensitivity: Hysteresis: Accuracy: AC coupled DC coupled Input hysteresis voltage Time base Accuracy: 207 Frequency, Period, Positive Width, Negative Width, Duty Cycle 3Hz to >125MHz 100mHz to >125MHz REF / COUNT (AC) IN TRIG / COUNT (DC) IN Up to 7 digits displayed. Automatic ≤50MHz 100mVpp - 5Vpp >50MHz 250mVpp - 5Vpp Maximum input ±10V Threshold typically 1.2V; Sensitivity 100mVpp (≤50MHz), 250mVpp (<50MHz) maximum input +5V, -1V 10mV ±1 digit ± time base accuracy. <± 1ppm initial settling error, <± 1ppm oscillator ageing rate in the first year, <1ppm over the specified temperature range TGF4000 Series Quick Start Guide 7 - Specification MODEL: TGF4042 TGF4082 TGF4162 TGF4242 Outputs Main Outputs Output Impedance: 50Ω 20mVp-p to 20Vp-p open circuit, 10mVp-p to 10Vp-p into 50Ω 20mVp-p to 20Vp-p open circuit, 10mVp-p to 10Vp-p into 50Ω 20mVp-p to 10Vp-p open circuit, 10mVp-p to 5Vp-p into 50Ω ≤ 50MHz Amplitude (Sine): ≤ 80MHz 20mVp-p to 20Vp-p open circuit, 10mVp-p to 10Vp-p into 50Ω 20mVp-p to 10Vp-p open circuit, 10mVp-p to 5Vp-p into 50Ω 20mVp-p to 5Vp-p open circuit, 10mVp-p to 2.5Vp-p into 50Ω ≤ 120MHz ≤ 240MHz 20mVp-p to 20Vp-p open circuit, 10mVp-p to 10Vp-p into 50Ω Amplitude (Pulse): ≤ 50MHz ≤ 100MHz 20mVp-p to 20Vp-p open circuit, 10mVp-p to 10Vp-p into 50Ω 20mVp-p to 8Vp-p open circuit, 10mVp-p to 4Vpp into 50Ω Amplitude can be specified open circuit (hi Z) or into an assumed load of 1Ω to 10kΩ in Vpp Amplitude Accuracy: 1.5% ±5mV at 1kHz into 50Ω DC Offset Range: ±10V. DC offset plus signal peak limited to ±10V from 50Ω. DC Offset Accuracy: Typically 1% ±50mV. Resolution: 3 digits or 1mV for both Amplitude and DC Offset. Sync Output Channel 2 can be configured to output Channel 1 sync from its MAIN OUT 2 socket. Sync is a multi-function output which is automatically selected to be any of the following. Alternatively, the user can choose Sync to always be carrier referenced, to output the currently used trigger signal or turn it off. ≤ 28.125MHz ≤ 62.5MHz A square wave with 50% duty cycle at the A square wave with 50% duty cycle at the waveform frequency. waveform frequency. Sine / Square / Ramp / Pulse / Arbs Carrier Waveform ≤ 80MHz ≤ 240MHz Sync: A sine wave at the waveform frequency. A sine wave at the waveform frequency. Pattern A positive pulse which is 1 bit rate wide at the beginning of the sequence Noise No sync associated with noise. A square wave with 50% duty cycle referenced to the internal modulation waveform when modulation source is internal, or a square wave referenced to the carrier waveform when AM/FM/PM/SUM/ modulation source is external. No sync is associated with Noise and DC waveforms as the modulation PWM source. A square wave referenced to the trigger rate. The sync is a TTL high when hop amplitude is the ASK output amplitude and TTL low when carrier amplitude is the output amplitude for positive slope and vice versa for negative slope. Modulation Sync: A square wave referenced to the trigger rate. The sync is a TTL high when hop frequency is the output FSK frequency and TTL low when carrier frequency is the output frequency for positive slope and vice versa for negative slope. A square wave referenced to the trigger rate. The sync is a TTL high when the hop phase is the output BPSK phase and TTL low when carrier phase is the output phase for positive slope and vice versa for negative slope. A square wave that is a TTL high from the beginning of the sweep and a TTL low from the midpoint of Sweep Sync: the sweep Internal Trigger A square wave with 50% duty cycle at the trigger frequency. Burst Sync: External Trigger A square wave with same duty cycle and frequency as the external source. Manual Trigger A positive pulse which is approximately 18us wide at the beginning of the event. Trigger: Selects the current trigger signal Output Signal Level: Logic level nominally 3V Output Impedance: 50Ω 208 TGF4000 Series Quick Start Guide 7 - Specification MODEL: TGF4042 TGF4082 TGF4162 TGF4242 Ref Clock Output Buffered version of the 10MHz clock currently in use (internal or external) Output Level: Nominally 3V logic level from 50Ω. Inputs Trigger / Count (DC) Input For ASK, FSK, BPSK, triggered sweep, gated burst, triggered burst and DC coupled external frequency measurement. Trigger Input DC – 1MHz Frequency Range: Counter Input 100mHz to >125MHz Threshold typically 1.2V; Signal Range: Sensitivity 100mVpp (≤50MHz), 250mVpp (>50MHz) Maximum input +5V / -1V. Minimum Pulse Width (Trigger Input): 50ns Polarity (Trigger Input): Selectable as high/rising edge or low/falling edge. Input Impedance: 10kΩ External Modulation Input For AM, FM, PM, SUM and PWM Voltage Range: Input Impedance: Bandwidth: Ref Clock / Count (AC) Input ± 2.5V full scale 5kΩ typical DC to 5MHz Input for an external 10MHz reference clock and AC coupled external frequency measurement. ≤50MHz 100mVpp – 5Vpp Voltage Range: >50MHz 250mVpp – 5Vpp Maximum Voltage: +10V Minimum Voltage: -10V Interfaces Full digital remote control facilities are available through LAN, USB and optional GPIB interfaces. LAN Interface Ethernet 100/10base – T hardware connection. 1.5 LXI Device Specification 2016 USB Interface Standard USB 2.0 hardware connection. Implemented as virtual-COM port. USB Flash Drive For waveform and set-up storage/recall. GPIB (optional) Conforming with IEEE488.1 and IEEE488.2 General 4.3 inch (10.9 cm) transflective backlit TFT LCD, 480 x 272 pixels, 262144 colours, adjustable brightness and contrast. Data Entry: Keyboard selection of mode, waveform etc.; value entry direct by numeric keys or by rotary control. Up to 9 complete instrument set−ups may be stored and recalled from Stored Settings: internal memory. Bench Top: 97mm height; 250mm width; 295mm depth Size: Rack mount: 86.5mm (2U) height; 213.5mm (½−rack) width; 269mm depth Weight: 3.2kg Power: 100-240VAC ±10% 50/60Hz ; 100-120VAC ±10% 400Hz ; 60VA max. Installation Category II. Operating Range: +5°C to 40°C, 20−80% RH. Storage Range: −20°C to + 60°C. Environmental: Indoor use at altitudes up to 2000m, Pollution Degree 2. Options: 19 inch rack mounting kit. Complies with EN61010−1 & EN61326-1. Safety & EMC: For details, request the EU Declaration of Conformity for this instrument via http://www.aimtti.com/support (serial no. needed). For details, request the EU Declaration of Conformity for this instrument via http://www.aimtti.com/support (serial no. needed). General specifications apply for the temperature range 5°C to 40°C. Accuracy specifications apply for the temperature range 18°C to 28°C after 30 minutes warm-up, at maximum output into 50Ω. Typical specifications are determined by design and are not guaranteed. Display: 209 TGF4000 Series Quick Start Guide EXCELLENCE THROUGH EXPERIENCE Aim-TTi is the trading name of Thurlby Thandar Instruments Ltd. (TTi), one of Europe’s leading manufacturers of test and measurement instruments. The company has wide experience in the design and manufacture of advanced test instruments and power supplies built up over more than thirty years. The company is based in the United Kingdom, and all products are built at the main facility in Huntingdon, close to the famous university city of Cambridge. TRACEABLE QUALITY SYSTEMS TTi is an ISO9001 registered company operating fully traceable quality systems for all processes from design through to final calibration. ISO9001:2015 Certificate number FM 20695 WHERE TO BUY AIM-TTI PRODUCTS Aim-TTi products are widely available from a network of distributors and agents in more than sixty countries across the world. To find your local distributor, please visit our website which provides full contact details. Designed and built in Europe by: Thurlby Thandar Instruments Ltd. Glebe Road, Huntingdon, Cambridgeshire. PE29 7DR United Kingdom Tel: +44 (0)1480 412451 Fax: +44 (0)1480 450409 Email: [email protected] Web: www.aimtti.com *48591-10* *01*
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211

Aim-TTI TGF4162 Guía de inicio rápido

Tipo
Guía de inicio rápido
Este manual también es adecuado para