Hach SD900 Basic User Manual

Tipo
Basic User Manual
DOC026.97.80437
SD900 Portable Sampler
02/2015, Edition 2
Basic User Manual
Manuel d'utilisation de base
Manual básico del usuario
Manual Básico do Usuário
基本用户手册
基本取扱説明書
기본 사용 설명서
English..............................................................................................................................3
Français......................................................................................................................... 28
Español.......................................................................................................................... 54
Português...................................................................................................................... 80
中文............................................................................................................................... 106
日本語........................................................................................................................... 129
한글............................................................................................................................... 153
2
Table of contents
Specifications on page 3 Operation on page 17
General information on page 5 Maintenance on page 23
Installation on page 8
Additional information
Additional information is available on the manufacturer's website.
Specifications
Specifications are subject to change without notice.
SD900 Portable Sampler
Specification Standard base Compact base Composite base
Dimensions Diameter: 50.5 cm (19.9 in.)
Height: 69.4 cm (27.3 in.)
Diameter: 44.1 cm (17.4 in.)
Height: 61 cm (24 in.)
Diameter: 50.28 cm
(19.8 in.)
Height: 79.75 cm
(31.4 in.)
Weight 15 kg (35.6 lb) with 1-L
polyethylene bottles (24x)
14.8 kg (32.6 lb) with 10-L (2.5 gal)
polyethylene container (1x)
12.2 kg (27 lb) with 575-mL
(19.44 oz) polyethylene
bottles (24x)
12.9 kg (28.3 lb) with 10-L
(2.5 gal) polyethylene
container (1x)
15 kg (36 lb) with 950-
mL (32.12 oz) glass
bottles (12x)
Enclosure Impact-resistant ABS, 3-section construction; double-walled base with 2.54 cm (1 in.)
insulation—direct bottle contact with ice.
Sample
temperature
0–60 °C (32–140 °F)
Strainers 316 stainless steel in standard size, high velocity or low profile for shallow depth applications
and Teflon
®
or 316 stainless steel in standard size
Sample intake
tubing
9.5 mm (3/8 in.) I.D. vinyl or Teflon-lined polyethylene
Sample bottle
capacity
1-L (0.26 gal) polyethylene and/or
350-mL (11.83 oz) glass bottles
(24x)
2.3-L (0.6 gal) polyethylene and/or
1.9-L (0.5 gal) glass bottles (8x)
3.8-L (1 gal) polyethylene and/or
3.8-L (1 gal) glass bottles (4x)
3.8-L (1 gal) polyethylene and/or
3.8-L (1 gal) glass bottles (2x)
21-L (5.5 gal) or 15-L (4 gal)
polyethylene composite container
or 20-L (5.25 gal) polyethylene or
10-L (2.5 gal) polyethylene or 10-L
(2.5 gal) glass (1x)
575-mL (19.44 oz)
polyethylene bottles (24x)
950-mL (32.12 oz) glass
bottles (8x)
10-L (2.5 gal) polyethylene
bottle (1x)
10-L (2.5 gal) glass bottle
(1x)
21-L (5.5 gal)
polyethylene bottle (1x)
English 3
SD controller
Specification Details
Dimensions (W x H x D) 29.2 x 17.1 x 26.4 cm (11½ x 6¾ x 10
3
/
8
in.)
Weight 4.2 kg (9.26 lb)
Enclosure PC/ABS blend, NEMA 4X, 6, IP 67
Power requirements 15 VDC supplied by a 8754500 power supply; 15 VDC supplied by an integral
power supply
Overload protection 7 A, DC line fuse for the pump
Operating temperature 0 to 50 °C (32 to 122 °F)
Storage temperature –30 to 60 °C (–22 to 140 °F)
Storage/operating humidity 100% condensing
Pump Peristaltic high speed, with spring-mounted Nylatron rollers
Pump enclosure IP37
Pump tubing 9.5 mm ID x 15.9 OD mm (
3
/
8
in. ID x
5
/
8
in. OD) silicone
Pump tubing life 20,000 sample cycles with: 1 L (0.3 gal) sample volume, 1 rinse, 6 minute
pacing interval, 4.9 m (16 ft) of
3
/
8
in. intake tube, 4.6 m (15 ft) of vertical lift,
21 °C (70 °F) sample temperature
Vertical sample lift Maximum 8.5 m (28 ft) for: 8.8 m (29 ft) of
3
/
8
-in. vinyl intake tube at sea level
at 20–25 °C (68–77 °F)
Pump flow rate 4.8 L/min (1.25 gpm) at 1 m (3 ft) vertical lift with
3
/
8
-in. intake tube typical
Sample volume Programmable in 10-mL (0.34 oz) increments from 10 to 10,000 mL (3.38 oz
to 2.6 gal)
Sample volume repeatability
(typical)
±5% of 200 mL sample volume using uncalibrated liquid detect with: 4.6 m
(15 ft) vertical lift, 4.9 m (16 ft) of
3
/
8
-in. vinyl intake tube, single bottle, full
bottle shut-off at room temperature and 1524 m (5000 ft) elevation
Sample volume accuracy
(typical)
±10% of 200 mL sample volume using uncalibrated liquid detect with: 4.6 m
(15 ft) vertical lift, 4.9 m (16 ft) of
3
/
8
-in. vinyl intake tube, single bottle, full
bottle shut-off at room temperature and 1524 m (5000 ft) elevation
Sampling modes Pacing: Time-fixed, flow-fixed, time-variable, flow-variable, constant time
variable volume (CTVV).
Refer to Constant Time Variable Volume (CTVV) sampling in the expanded
version of this manual.
Distribution: Single bottle composite, multi-bottle composite, multi-bottle
discrete, bottles per sample, samples per bottle, combination of bottles per
sample, samples per bottle
Run modes Continuous or non-continuous with user-entered number of samples
Multiple programs Stores up to three sampling programs
Transfer velocity (typical) 0.9 m/s (2.9 ft/s) with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of
3
/
8
-in. vinyl
intake tubing, 21 °C (70 °F) and 1524 m (5000 ft) elevation
Liquid sensor Ultrasonic. Body: Ultem
®
NSF ANSI standard 51 approved, USP Class VI
compliant
Sample history Up to 510 records
Air purge Air purged automatically before and after each sample. The duration
automatically compensates for varying intake tube lengths.
Event log 510 records
4 English
Specification Details
Connections Power, auxiliary, serial communications, distributor, SDI-12
Wetted materials Stainless steel, polyethylene, Teflon, Ultem, silicone
Communications RS232, Modbus, SDI-12
Warranty 1 year
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
English 5
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential
injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates a potential pinch hazard.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public
disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to
the user.
Certification
Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing
Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC
Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including interference that may cause
undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment. This equipment has been tested
and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case the user will be required to
correct the interference at their expense. The following techniques can be used to reduce
interference problems:
1. Disconnect the equipment from its power source to verify that it is or is not the source of the
interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect
the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.
6
English
Product overview
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or
chemical feed system for which there are regulatory limits and monitoring requirements related to
public health, public safety, food or beverage manufacture or processing, it is the responsibility of the
user of this instrument to know and abide by any applicable regulation and to have sufficient and
appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction
of the instrument.
C A U T I O N
Fire hazard. This product is not designed for use with flammable liquids.
The SD900 Portable Sampler collects liquid samples at specified intervals and keeps the samples in
bottles or containers. Use the sampler for a wide variety of aqueous applications and for toxic
pollutants and suspended solids. Set up the sampler with different retainers, bottles or containers.
Refer to Figure 1.
Figure 1 Product overview
1 Compact base 6 Controller connections
2 Standard insulated base 7 Pump
3 Center section 8 Controller
4 Power source 9 Liquid sensor
5 Top cover 10 Standard insulated base for 21-L (5.5 gal)
container
Product components
Make sure that all components have been received. Refer to Figure 2. If any items are missing or
damaged, contact the manufacturer or a sales representative immediately.
English
7
Figure 2 Product components
1 Base (Standard, compact or composite) 6 Battery charger (optional)
2 Components for a single-bottle option (bottle and
support can change)
7 AC power supply (optional)
3 Center section with controller 8 Strainer
4 Battery (optional) 9 Intake tubing, vinyl or Teflon-lined
5 Top cover 10 Components for a multiple-bottle option (bottles
and retainers can change)
Installation
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Confined space precautions
D A N G E R
Explosion hazard. Training in pre-entry testing, ventilation, entry procedures, evacuation/rescue
procedures and safety work practices is necessary before entering confined spaces.
The information that follows is supplied to help users understand the dangers and risks that are
associated with entry into confined spaces.
On April 15, 1993, OSHA's final ruling on CFR 1910.146, Permit Required Confined Spaces, became
law. This standard directly affects more than 250,000 industrial sites in the United States and was
created to protect the health and safety of workers in confined spaces.
Definition of a confined space:
A confined space is any location or enclosure that has (or has the immediate potential for) one or
more of the following conditions:
An atmosphere with an oxygen concentration that is less than 19.5% or more than 23.5% and/or a
hydrogen sulfide (H
2
S) concentration that is more than 10 ppm.
An atmosphere that can be flammable or explosive due to gases, vapors, mists, dusts or fibers.
Toxic materials which upon contact or inhalation can cause injury, impairment of health or death.
8
English
Confined spaces are not designed for human occupancy. Confined spaces have a restricted entry
and contain known or potential hazards. Examples of confined spaces include manholes, stacks,
pipes, vats, switch vaults and other similar locations.
Standard safety procedures must always be obeyed before entry into confined spaces and/or
locations where hazardous gases, vapors, mists, dusts or fibers can be present. Before entry into a
confined space, find and read all procedures that are related to confined space entry.
Mechanical installation
Site installation guidelines
D A N G E R
Explosion hazard. The instrument is not approved for installation in hazardous locations.
Refer to the guidelines that follow for the site location evaluation.
Obey all the safety precautions if the sampler is installed in a confined space. Refer to Confined
space precautions on page 8.
Make sure that the temperature at the location is in the specification range. Refer to Specifications
on page 3.
Install the sampler on a level surface or hang the sampler with the suspension harness, the
support bracket or the spanner bar. Refer to Install the sampler in a manhole on page 9 and to
the applicable installation documentation.
As near the sample source as possible to decrease analysis delay. Refer to Plumb the sampler
on page 12.
For limitations on transport velocity and maximum vertical lift, refer to Specifications on page 3.
Install the sampler in a manhole
Install the sampler above the sample water in a manhole. Install the sampler with a spanner bar or a
support bracket. Install the spanner bar inside the manhole. The spanner bar is supported by
pressure against the walls. The support bracket has the same width as the manhole cover. Install the
support bracket directly below the cover for support.Refer to the expanded manual on the
manufacturer's website.Refer to the documentation supplied with the accessories to install the
sampler.
Install the distributor or full-bottle shutoff (optional)
The distributor or full-bottle shutoff assembly is typically installed at the factory. Refer to the
distributor or full-bottle shutoff documentation for installation.
Prepare the sampler
Clean the sample bottles
Clean the sample bottles and caps with a brush, water and a mild detergent. Flush the containers
with fresh water followed by a distilled water rinse.
Install a single bottle
Use a single bottle to collect one composite sample. When the bottle is full, the full bottle shut-off
stops the sample program.
1. Clean the sample bottles. Refer to Clean the sample bottles on page 9.
2. Install the sample bottle as shown in Figure 3.
English
9
Figure 3 Single bottle installation
1 Polyethylene bottle, 10 L (2.5 gal) (1918) 6 Polyethylene bottle, 21 L (5.5 gal) (6494)
2 Glass bottle, 10 L (2.5 gal) (6559) 7 Compact base (8975)
3 Support (1502) 8 Standard insulated base (8976)
4 Polyethylene bottle, 15 L (4 gal) (1367) 9 Standard insulated base for 21 L (5.5 gal) bottle
(8561)
5 Polyethylene bottle, 19 L (5 gal) (6498)
Install the first bottle for multiple sample collections
Use multiple bottles to collect samples into separate bottles or into more than one bottle. The
distributor moves the sample tube above each bottle. Install the bottles in the sampler base as shown
in Figure 4. Install the first sample bottle (number 1) below the label in the sampler base. Install the
remaining bottles in increasing numbers in the direction shown by the label. Refer to Figure 5
on page 12 for a diagram of necessary components.
10
English
Figure 4 Bottle number 1 installation
1 Bottle number 1 location (compact base) 6 Elastic straps
2 Bottle number 1 location for 24 bottles 7 Standard insulated base for 21 L (5.5 gal) bottle
3 Bottle number 1 location for 8 bottles 8 Standad insulated base
4 Bottle number 1 location for 2 or 4 bottles 9 Compact base
5 Retainer
Install multiple bottles
When multiple bottles are installed, a distributor arm moves the sample tube over each bottle.
Sample collection automatically stops when the specified number of samples have been collected.
1. Clean the sample bottles. Refer to Clean the sample bottles on page 9.
2. Install the first sample bottle (number 1) below the label in the sampler base. Refer to Install the
first bottle for multiple sample collections on page 10.
3. Assemble the sample bottles as shown in Figure 5. For eight or more bottles, make sure that the
first bottle goes next to the bottle one indicator in the clockwise direction.
4. Put the bottle assembly in the sampler. For eight or more bottles, align the wires in the slots in the
bottom tray.
English
11
Figure 5 Multiple bottle installation
1 Positioner/retainer (2347) 8 Glass bottles, 3.8 L (1 gal), (4x) (2216) or
polyethylene bottles, 3.8 L (1 gal), (4x) (2217)
2 Retainer (1422) 9 Glass bottles, 1.9 L (0.5 gal), (8x) (1118) or
polyethylene bottles, 2.3 L (0.61 gal), (8x) (657)
3 Retainer (2190) 10 Polyethylene bottles, 1 L (0.26 gal), (24x) (737)
4 Retainer (2189) 11 Glass bottles, 350 mL (11.83 oz), (24x) (732)
5 Glass bottles, 950 mL (32.12 oz) (8x) (2348) 12 Compact base (8975)
6 Polyethylene bottles, 575 mL (19.44 oz), (24x)
(1369)
13 Standard base (8976)
7 Glass bottles, 3.8 L (1 gal), (2x) (2214) or
polyethylene bottles, 3.8 L (1 gal), (2x) (2215)
Plumb the sampler
Install the strainer in the middle of the sample stream (not near the surface or the bottom) to make
sure that a representative sample is collected. Refer to Figure 6. Refer to Figure 7 for the intake tube
installation.
1. Connect the tubing to the sampler as shown in Figure 7.
Note: Use connection kit 2186 if Teflon-lined tubing is used.
2. Install the intake tube and strainer in the main stream of the sample source where the water is
turbulent and well-mixed.
Make the intake tube as short as possible.
Keep the intake tube at a maximum vertical slope so that the tube drains completely between
samples.
Note: If a vertical slope is not possible or if the tube is pressurized, disable the liquid sensor. Calibrate the
sample volume manually. Refer to Calibrate the sample volume manually on page 22.
Make sure that the intake tube is not pinched.
12
English
Figure 6 Instrument installation
1 Strainer 3 Intake tube
2 Vertical lift
Figure 7 Intake tube installation
Electrical installation
Controller connections
W A R N I N G
Electrical shock hazard. Externally connected equipment must have an applicable country safety
standard assessment.
Figure 8 shows the connections that can be made to the controller.
English
13
Figure 8 Controller connectors
1 Auxiliary device 4 SDI-12 device option
2 Power supply 5 Distributor/full bottle shut-off
3 Serial communications
Connect a flow meter
Connect a flow meter to the controller to start or stop the sampler when the sample flow goes above
or below a specified value.
Items to collect:
Multi-purpose full cable for Sigma flow meters (or 980 half cable for the model 980 flow meter).
Optional splitter for additional connections. Two or more splitters can be connected in series.
1. Connect one end of the cable to the flow meter. For the model 980 flow meter, refer to the model
980 flow meter user manual.
2. Connect the other end of the cable to the auxiliary device connector on the controller.
Note: If the flow meter has a 6-pin cable, use the 6-pin to 7-pin adapter cable.
Connect a PC or communications network
Connect a PC or a communications network to the controller to transfer data or to configure the
sampler.
Items to collect:
Serial cable, 7-pin RS232 to DB-9 (refer to Figure 9). Connections: B to 5 (signal ground); D to
3 (RCD); F to 2 (TXD); G (ground).
1. Connect one end of the serial cable to the communications device or network.
2. Connect the other end of the cable to the serial communications connector on the controller.
Figure 9 Serial communications cable
14 English
Connect the sampler to power
D A N G E R
Electrocution hazard. If this equipment is used outdoors or in potentially wet locations, a Ground Fault
Circuit Interrupt (GFCI/GFI) device must be used for connecting the equipment to its main power
source.
D A N G E R
Fire hazard. Install a 15 A circuit breaker in the power line. A circuit breaker can be the local power
disconnect, if located in close proximity to the equipment.
D A N G E R
Electrocution hazard. Protective Earth Ground (PE) connection is required.
W A R N I N G
Electrocution hazard. Make sure that there is easy access to the local power disconnect.
W A R N I N G
Electric shock hazard. The power supply can overheat when the time between pump cycles is too
short.
N O T I C E
Make sure that the pump cycle time will not cause the power supply to overheat. Refer to Calculate the pump
cycle time on page 16.
Connect the sampler to a battery or to an AC power supply. Refer to the illustrated steps that follow.
English
15
Calculate the pump cycle time
The pump must stay off for a minimum time period between sample cycles so that the power supply
does not overheat.
1. Calculate the total amount of time that the pump will run continuously during a sample cycle.
Include all stages: pre-purge, intake rinse, sample, sample retries and post-purge.
2. Find the minimum amount of time that the pump must stay off between sample cycles from
Figure 10.
3. Make sure that the pacing or time interval in the sampler program lets the pump stay off between
sample cycles for the minimum amount of time.
Figure 10 Pump on and off times at 50 °C (122 ºF)
1 Time (in minutes) that the pump must stay off 3 Example: if the pump is on continuously for
5 minutes, the pump must stay off for 15 minutes.
2 Time (in minutes) that the pump is on continuously
Set the power to on
Push the power button to set the power to on or off. Make sure that the power supply (AC power or
battery power) is correctly installed.
User interface and navigation
The controller keypad is shown in Figure 11. Use the MENU key to set up sampling programs,
configure the controller settings or complete the diagnostic tests. Use the arrows, ENTER and BACK
keys to scroll through the menu, make selections and enter values. Look for arrows on the bottom or
side of the display screen to know when more options are available.
16
English
Figure 11 Controller keypad
1 POWER 5 MENU 9 STATUS
2 VOLUME CALIBRATION 6 ENTER 10 RUN/HALT PROGRAM
3 STOP 7 BACK 11 MANUAL OPERATION
4 LED 8 ARROW KEYS
Operation
Configure the system settings
1. Push MENU.
2. Go to SYSTEM SETUP.
3. Change the applicable options.
Option Description
Time/date setup Set the time (24-hour format) and date.
Communication Select the baud rate (19200, 38400, 57600 or 115200) and protocol (Modbus RTU or
ASCII) for the serial port.
Setup base Select the sampler base (All weather refrigerated, refrigerated, portable standard or
portable compact).
Language Select the language of the controller.
Liq sensor cal Calibrate the liquid sensor or set to default.
Set contrast Adjust the contrast of the LCD screen.
Pump duty cycle Change the pump duty cycle. Range: 50% to 100% (default: 100%).
Password setup Set up a password to control access to the program setup and the system setup menus.
Select ENABLE>YES to set a new password or ENABLE>NO to activate the current
password. If the password is set for the first time, use 900900 as the current password.
English 17
Option Description
Tubing life Set a reminder when the pump tubing is to be changed. Go to ENABLE>RESET
CYCLES to reset the pump cycle count to 0. Go to ENABLE>CYCLE LIMIT to change
the number of pump cycles that is used for the reminder. Go to ENABLE>STATUS to
show the current number of pump cycle counts.
SDI–12 (optional) Configure the operation of an SDI–12 device. This option is shown only when a device is
found.
Set up a sampling program
A sampling program includes all of the parameters that are necessary to collect samples
automatically. Collect the samples at regular intervals or when the sampling program is complete.
1. Push MENU.
2. Go to PROGRAM SETUP>MODIFY ALL. The first screen is shown.
3. Select or enter the values for each parameter. Use the left and right arrow keys to move left or
right. Use the up and down arrow keys to change a value. Change the units if necessary (gal or
mL) with the arrow keys.
Option Description
Bottles Bottle quantity: the number of bottles in the sampler (1, 2, 4, 8 or 24).
Full BTL shutoff: Enabled/Disabled. If enabled, the controller checks for bottle
overflow conditions.
Bottle volume: the volume capacity of each bottle. Range: 50–65000 mL (0.01 -
17.17 gal)
Intake tubing Tubing length: the length of the intake tubing from the strainer to the liquid sensor.
Range: 100–3000 cm or 3–99 ft. An accurate length is necessary to get an accurate
sample volume.
Intake tube type: the diameter and material of the intake tubing (¼ in. vinyl, 3/8 in.
vinyl or 3/8 in. Teflon).
Program delay Enable/disable: when enabled, the sampling program starts at a specified time or
number of counts. Date and time: the date and time when the program starts (24-
hour format). Counts: the number of counts or pulses from a flow meter when the
program will start (1 to 9999 counts). If the sample pacing is later set to time, the
program delay will be disabled.
Sample
pacing/collection
Time based—Refer to Time based collection on page 19.
Note: When CTVV is used:
The liquid sensor is always enabled.
The sample distribution is composite.
The run mode is non-continuous.
The first sample is taken after the interval.
Flow source is always 4–20 mA input.
Flow based—Refer to Flow based collection on page 20.
Sample distribution Deliver samples to all>YES
Each sample is delivered to all bottles. End after last: program stops after the last
sample (1–999). Continuous: program continues until it is stopped manually.
Deliver samples to all>NO
Samples are delivered to a subset of bottles . Samples/bottle: the number of samples
to be collected in each bottle (1–999). Bottles/sample: The number of bottles that will
contain the same sample.
Liquid sensor Enable or disabled. If disabled, the sample volume must be calibrated by time.
18 English
Option Description
Sample volume The volume of each sample in mL (1 gal = 3785.4 mL). If bottles/sample mode is
used, each bottle receives a full sample volume (10–10,000 mL). Make sure that the
sample volume does not exceed the bottle volume. Sample volumes are rounded to
the nearest 10 mL.
Intake rinses The number of intake tube rinses before a sample is collected (0–3).
Sample retries The number of sampling tries after a sampling failure occurs (0–3).
Site ID The name for the sampling location (up to 12 characters). The site ID is used as the
preset name if the program is saved.
Advanced sampling Goes to the advanced sampling menu. If necessary, configure the options in the
Advanced Sampling menu.
Run program The creation of the basic sampling program finishes. The sampler prompts the user
to start or cancel the program.
Time based collection
1. Select or enter the values for each option.
Option Description
Pacing interval Collect samples at regular time intervals, in hours and minutes (0:01–999:00).
Take first sample The program starts immediately or after the first interval.
Variable volume Specify whether the sample volumes vary (flow based). If yes, refer to Use variable
volume on page 19. If no, refer to Do not use variable volume on page 20.
Use variable volume
Sample pacing is constant time variable volume (CTVV).
1. Select or enter the values for each option.
Option Description
Select unit Select the flow units to use.
Map 4 mA input Enter the flow rate that agrees with 4 mA input.
Map 20 mA input Enter the flow rate that agrees with 20 mA input.
Average flow rate Specify the average flow rate for the site (1–999,999). The units are specified by
the units selected above.
Time volume desired Specify the total volume to be collected during the sample program
(10–10,000 ml).
English 19
Option Description
Collection period Specify the total collection period in hours and minutes (0:01–999:00).
Minimum sample volume If the flow volume is not sufficient to collect this minimum volume, the sample is
skipped (10–10,000 ml).
Do not use variable volume
1. Select or enter the values for the option.
Option Description
Take first sample Select whether the program starts immediately or after the first time interval has passed.
Flow based collection
Specify the flow source for flow based sampling:
Counts—Refer to Use counts for the flow source on page 20.
4–20 mA input—Refer to Use 4–20 mA input for the flow source on page 20.
Use counts for the flow source
1. Select or enter the values for each option.
Option Description
Take sample every Collect samples at regular flow intervals, in counts (1–9999 counts).
Overrride time Collect a sample if the flow volume is very low.
Time The maximum time between samples (0:01–999:00). The timer starts again after each
sample is collected.
Take first sample The program starts immediately or after the first time interval.
Use 4–20 mA input for the flow source
1. Select or enter the values for each option.
Option Description
Select unit Selects the flow units to use.
Map 4 mA input Enter the flow rate that would agree with the 4 mA input.
Map 20 mA input Enter the flow rate that would agree with the 20 mA input.
Take sample every Enter the total flow after which the sample has to be drawn.
Override time Select enable to push a sample to be collected if the flow volume is unusually low.
Time The maximum time between samples (0:01–999:00). The timer starts again after each
sample is collected.
Sample volume Enter the volume to be collected per sample. If bottles/sample mode is selected, each
bottle receives a full sample volume (100–10,000 mL). This option is skipped if variable
volume is selected.
Restore the default settings
The restore option sets all of the program parameters to the default settings.
20
English
1. Push MENU.
2. Go to PROGRAM SETUP>RESTORE.
3. Select YES.
Start or stop a program
Use the RUN/HALT PROGRAM key to start or stop a program. Up to 12 user-defined start/stop
times/dates can be set. A program must be stopped before the program or system setup can be
changed.
1. Push RUN/HALT PROGRAM.
2. Select one of the options.
Option Description
Start Starts the currently loaded program. The status changes to RUNNING.
Halt Stops the program temporarily. The status changes to HALTED.
Resume The program resumes from the point at which it was halted. The status changes to
RUNNING.
Start over The program starts from the beginning. The status changes to RUNNING.
End program Stops the currently running program. The status changes to COMPLETE.
Manual operation
Use manual operation to collect a grab sample, move the distributor arm or operate the pump. For
additional information, refer to the expanded version of this manual.
Volume calibration
The sample volume can be calibrated automatically with the liquid sensor or manually. When the
liquid sensor is used for volume calibration, the sensor is adjusted to accurately measure all
programmed volumes. If the liquid sensor is disabled, the volume for samples, rinses and stormwater
must be calibrated manually.
For typical applications, tap water can be used for the volume calibration. If the sample composition
is significantly different from typical water samples, calibrate the sampler with the liquid to be
collected. All programs must be stopped before calibration.
Calibrate the liquid sensor
The accuracy of the liquid sensor varies with the type of liquid that is sampled. For example, the
volume of a highly turbid sample may not be as accurate as that of clear water. The liquid sensor can
be calibrated with the sample water to improve volume accuracy.
1. Put the intake tubing into the sample water.
2. Push MENU.
3. Go to PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE to make sure that
the liquid sensor is enabled.
4. Go to SYSTEM SETUP>LIQ SENSOR CAL.
5. Select PERFORM CAL.
6. Select START. The pump operates in reverse to purge the intake tube. The pump then operates
in the forward direction.
7. Look at the sample flow in the intake tube to make sure that the sample goes through the liquid
sensor.
8. When the sample goes through the liquid sensor and no bubbles are present, push STOP. The
intake tube is automatically purged. The liquid sensor calibration is complete.
9. Select DONE to exit or REPEAT to do the calibration again.
English
21
Restore the default calibration
The volume calibration for the liquid sensor can be returned to the default factory settings.
1. Select PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE to make sure that
the liquid sensor is enabled.
2. Push VOLUME CALIBRATION and select RESET CAL. The volume calibration is set to the
default settings. The message “value has been reset to 0” will be shown.
Calibrate the sample volume with the liquid sensor
Calibrate the sample volume with the liquid sensor enabled to adjust sample volumes slightly.
1. Go to PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE to make sure that
the liquid sensor is enabled.
2. Put the intake tubing into the sample water or tap water.
Note: Use the sample water for best accuracy.
3. Put the tubing from the outlet side of the pump into a graduated cylinder.
4. Push VOLUME CALIBRATION and select CALIBRATION.
5. Select START. The pump operates in reverse to purge the intake tube. The pump then operates
in the forward direction and collects the sample volume that is specified in the sampling program.
The pump operates in reverse to purge the intake tube.
6. When the sample has been collected, select DONE. Compare the volume of the sample in the
graduated cylinder with the sample volume that is entered in the program setup menu.
7. If the volume that was collected is different from the sample volume in the current program, enter
the volume that was actually collected. Push ENTER. The sensor is adjusted to accurately
measure all programmed volumes.
8. Select START to start the sampling program or CANCEL to exit the volume calibration menu.
Calibrate the sample volume manually
When the liquid sensor is disabled, all volumes must be calibrated manually by time. The sample
volume is calibrated for the volume specified in the current program. If the sample volume is changed
in the program, the sample volume must be calibrated again for the new volume.
1. Select PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>DISABLE to make sure that
the liquid sensor is disabled.
2. Put the intake tubing into the sample water or tap water.
3. Put the tubing from the outlet side of the pump into a graduated cylinder.
4. Push VOLUME CALIBRATION. If more than one volume is shown, select one of the volumes for
calibration.
5. Select START. The pump operates in reverse to purge the intake tube. The pump then operates
in the forward direction and starts to collect the sample.
6. Stop the pump at the selected volume or rinse location:
Sample volume: push STOP when the volume that is specified in the main program is
collected.
Rinse: push STOP when the liquid in the intake tubing gets to the liquid detector.
Storm volume: push STOP when the volume specified in the stormwater program is collected.
7. Select DONE to go to the next volume calibration or REPEAT to do the calibration again.
8. When all volumes have been calibrated, select DONE.
9. Connect the outlet tubing to the tube fitting on the sampler.
10. Select START to start the sampling program or CANCEL to exit.
22
English
Verify the sample volume
To make sure that the sample volume is accurate, take a grab sample. Do not go back into
calibration to check the volume since the volume compensation is reset to zero at the start of a
calibration.
1. Push MANUAL OPERATION.
2. Select GRAB SAMPLE.
3. Put the intake tubing into the sample water.
4. Put the tubing from the outlet side of the pump into a graduated cylinder.
5. If the liquid sensor is enabled, enter the volume to be verified.
6. If the liquid sensor is disabled, enter the volume that is specified in the program.
7. Push ENTER. The pump cycle starts (purge-sample-purge).
8. Compare the volume that was collected in the graduated cylinder to the expected volume. If the
collected volume is not correct, do the volume calibration again.
Maintenance
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
D A N G E R
Electrocution hazard. Remove power from the instrument before doing maintenance or service
activities.
W A R N I N G
Biohazard exposure. Obey safe handling protocols during contact with sample bottles and sampler
components.
W A R N I N G
Multiple hazards. The technician must make sure that the equipment operates safely and correctly
after maintenance procedures.
Clean the instrument
C A U T I O N
Fire hazard. Do not use flammable agents to clean the instrument.
N O T I C E
Do not clean the controller compartment heater with liquids of any kind.
If water is not sufficient to clean the controller and the pump, disconnect the controller and move the
controller away from the sampler. Allow sufficient time for the controller and pump to dry before the
parts are re-installed and put back into service.
Clean the sampler as follows:
Sampler enclosure—Clean the internal and external surfaces with a damp cloth and a mild
detergent. Do not use abrasive cleaners or solvents.
English
23
Sample bottles—Clean the bottles with a brush and water and a mild detergent. Rinse with fresh
water and then with distilled water. Use an autoclave to clean glass bottles if necessary.
Replace the desiccant
A desiccant cartridge in the controller absorbs moisture and prevents corrosion. Monitor the
desiccant color through the window (Figure 12). Fresh desiccant is yellow. When the color is green,
replace the desiccant.
1. Unscrew and remove the desiccant cartridge (Figure 12).
2. Remove the plug and discard the spent desiccant.
3. Fill the desiccant tube with fresh desiccant.
4. Install the plug.
5. Apply grease to the O-ring.
6. Install the desiccant tube in the controller.
Figure 12 Desiccant cartridge
1 Plug 3 O-ring
2 Desiccant tube 4 Desiccant window
Pump maintenance
C A U T I O N
Pinch hazard. Remove power from the instrument before maintenance or service activities are done.
Replace the pump tubing
N O T I C E
Use of tubing other than that supplied by the manufacturer can cause excessive wear on mechanical parts and/or
poor pump performance.
Examine the pump tubing for wear where the rollers rub against the tubing. Replace the tubing when
the reminder message is shown or when the tubing shows signs of wear.
Pre-requisites:
Pump tubing—pre-cut or bulk 4.6 m or 15.2 m (15 ft or 50 ft)
1. Disconnect the power to the controller.
2. If the bulk tubing is used, cut the tubing and add alignment dots. Refer to the expanded manual
on the manufacturer's website.
3. Remove the pump tubing as shown in the illustrated steps that follow.
4. Clean the silicone residue from the interior of the pump housing and from the rollers.
5. Install the new pump tubing as shown in the illustrated steps that follow.
24
English
English 25
Clean the rotor
Clean the rotor, pump tube tracks and pump housing with a mild detergent. Refer to Replace the
pump tubing on page 24 and the illustrated steps that follow.
Replace the distributor arm tube
The distributor arm moves over each bottle during multiple bottle sampling. Replace the tube in the
distributor arm when the tube is worn. Make sure that the correct tube is used for the correct
distributor and distributor arm.
Note: The distributor tubing is not the same as the pump tubing. The pump tubing installed in the distributor
assembly can damage the distributor. Also, samples can be missed because the distributor arm cannot move
easily.
1. Remove the tube from the distributor arm and from the ceiling of the center section.
2. Insert the new tube into the distributor arm. Make sure that the tube is aligned with the
dimensions shown in Figure 13.
3. Insert the other end of the tube into the fitting on the ceiling of the center section.
4. Complete the diagnostic test for the distributor to make sure that the operation is correct.
Figure 13 Distributor assembly
1 Nozzle 3 Distributor motor 5 Distributor tube
2 Distributor arm 4 Shaft
26 English
Fuse replacement
The fuse is a not a user-serviceable item. Contact technical support. A blown fuse can be an
indication that the sampler has a problem and that service is necessary.
English 27
Table des matières
Caractéristiques à la page 28 Fonctionnement à la page 42
Généralités à la page 30 Maintenance à la page 49
Installation à la page 33
Informations supplémentaires
Des informations supplémentaires sont disponibles sur le site Web du fabricant.
Caractéristiques
Les caractéristiques techniques peuvent être modifiées sans préavis.
Echantillonneur portable SD900
Caractéristique Base standard Base compacte Base composite
Dimensions Diamètre : 50,5 cm (19,9 po)
Hauteur : 69,4 cm (27,3 po)
Diamètre : 44,1 cm
(17,4 po)
Hauteur : 61 cm (24 po)
Diamètre : 50,28 cm
(19,8 po)
Hauteur : 79,75 cm
(31,4 po)
Poids 15 kg (35,6 lb) avec des bouteilles
en polyéthylène de 1 l (24x)
14,8 kg (32,6 lb) avec un récipient
en polyéthylène de 10 l (2,5 gal)
(1x)
12,2 kg (27 lb) avec des
bouteilles en polyéthylène
de 575 ml (19,44 oz) (24x)
12,9 kg (28,3 lb) avec un
récipient en polyéthylène de
10 l (2,5 gal) (1x)
15 kg (36 lb) avec des
bouteilles en verre de
950 ml (32,12 oz) (12x)
Boîtier ABS résistant aux impacts, construction en 3 sections, base à double paroi avec isolation
de 2,54 cm (1 po) ; contact direct entre la bouteille et la glace.
Température de
l'échantillon
0 à 60 °C (32 à 140 °F)
Arrêtoirs Acier inoxydable 316 dans la taille standard, haute vitesse ou bas profil pour des
applications peu profondes et acier inoxydable 316 ou en Téflon
®
dans la taille standard
Tuyauterie
d'admission de
l'échantillon
9,5 mm (3/8 po) DI en vinyle ou polyéthylène doublé en Téflon
Capacité de la
bouteille
d'échantillon
Bouteilles de 1 l (0,26 gal) en
polyéthylène et/ou de 350 ml
(11,83 oz) en verre (24x)
Bouteilles de 2,3 l (0,6 gal) en
polyéthylène et/ou de 1,9 l (0,5 gal)
en verre (8x)
Bouteilles de 3,8 l (1 gal) en
polyéthylène et/ou de 3,8 l (1 gal)
en verre (4x)
Bouteilles de 3,8 l (1 gal) en
polyéthylène et/ou de 3,8 l (1 gal)
en verre (2x)
Récipient de 21 l (5,5 gal) ou de
15 l (4 gal) en polyéthylène
composite ou de 20 l (5,25 gal) en
polyéthylène, ou de 10 l (2,5 gal)
en polyéthylène, ou de 10 l
(2,5 gal) en verre (1x)
Bouteilles en polyéthylène
de 575 ml (19,44 oz) (24x)
Bouteilles en verre de
950 ml (32,12 oz) (8x)
Bouteilles en polyéthylène
de 10 l (2,5 gal) (1x)
Bouteilles en verre de 10 l
(2,5 gal) (1x)
Bouteilles en
polyéthylène de 21 l
(5,5 gal) (1x)
28 Français
Contrôleur SD
Caractéristique Détails
Dimensions (L x H x P) 29,2 x 17,1 x 26,4 cm (11½ x 6¾ x 10
3
/
8
po.)
Poids 4.2 kg (9.26 lb)
Boîtier Alliage PC/ABS, NEMA 4X, 6, IP 67
Alimentation requise 15 V c.c. fournis par une alimentation 8754500 ; 15 V c.c. fournis par une
alimentation complète
Protection contre les surcharges 7 A, fusible de ligne c.c. pour la pompe
Température de fonctionnement 0 à 50 °C
Température de stockage –30 à 60 °C (–22 à 140 °F)
Humidité de stockage/de
fonctionnement
100% de condensation
Pompe Grande vitesse péristaltique, avec rouleaux en Nylatron montés à ressort
Corps de pompe IP37
Tuyau de pompe 9,5 mm DI x 15,9 mm DE (
3
/
8
po. DI x
5
/
8
po. DE) en silicone
Durée de vie de la pompe 20 000 cycles d'échantillonnage avec : 1 l (0,3 gal) de volume d'échantillon,
1 rinçage, 6 minutes d'intervalle, 4,9 m (16 pieds) du tube d'admission
de
3
/
8
po., 4,6 m (15 pieds) de levée verticale, température d'échantillon de
21 °C (70 °F)
Levée verticale de l'échantillon Maximum 8,5 m (28 pieds) pour : 8,8 m (29 pieds) du tube d'admission en
vinyle de
3
/
8
po. au niveau de la mer à 20-25 °C (68-77 °C)
Débit de pompe 4,8 l/min (1,25 gpm) avec une levée verticale de 1 m (3 ft) avec un tube
d'admission de
3
/
8
po. typique
Sample volume (Volume
d’échantillon)
Programmable par incréments de 10 ml (0,34 oz) de 10 à 10 000 ml (3,38 oz à
2,6 gal)
Répétabilité du volume
d'échantillon (typique)
±5 % de volume d'échantillon de 200 ml par détection de liquide non étalonné
avec 4,6 m (15 pieds) de levée verticale, 4,9 m (16 pieds) du tube d'admission
en vinyle de
3
/
8
po., bouteille seule, fermeture de bouteille pleine à
température ambiante et avec une altitude de 1 524 m (5 000 pieds)
Précision du volume
d'échantillon (typique)
±10 % de volume d'échantillon de 200 ml par détection de liquide non étalonné
avec : 4,6 m (15 pieds) de levée verticale, 4,9 m (16 pieds) du tube
d'admission en vinyle de
3
/
8
po. bouteille seule, fermeture de bouteille pleine à
température ambiante et avec une altitude de 1 524 m (5 000 pieds)
Modes d'échantillonnage Rythme : durée fixe, débit fixe, durée variable, débit variable, volume variable
à intervalle fixe (CTVV).
Reportez-vous à la section sur l'échantillonnage de volume variable à
intervalle fixe (CTVV) dans la version complète de ce manuel.
Distribution : composite bouteille unique, composite multi-bouteilles, discrète
multi-bouteilles, bouteilles par échantillon, échantillons par bouteille,
combinaison de bouteilles par échantillon, échantillons par bouteille
Modes d'exécution Continu ou non continu avec nombre d'échantillons entré par l'utilisateur
Programmes multiples Enregistre jusqu'à trois programmes d'échantillonnage
Vitesse du transfert (typique) 0,9 m/s (2,9 pieds/s) avec : 4,6 m (15 pieds) de levée verticale, 4,9 m
(16 pieds) du tube d'admission en vinyle de
3
/
8
po., 21 °C (70 °F) et 1 524 m
(5 000 pieds) d'altitude
Capteur de liquide Ultrasonique. Corps : Ultem
®
agréé NSF ANSI norme 51, conforme USP
classe VI
Français 29
Caractéristique Détails
Historique d'échantillonnage Jusqu'à 510 enregistrements
Purge d'air Air purgé automatiquement avant et après chaque échantillon. La durée
compense automatiquement des variations des longueurs de tube
d'admission.
Journal événements 510 enregistrements
Connexions Alimentation, auxiliaire, communication série, distributeur, SDL-12
Matériaux immergés Acier inoxydable, polyéthylène, Teflon, Ultem, silicone
Communications RS232, Modbus, SDI-12
Garantie 1 an
Généralités
En aucun cas le constructeur ne saurait être responsable des dommages directs, indirects, spéciaux,
accessoires ou consécutifs résultant d'un défaut ou d'une omission dans ce manuel. Le constructeur
se réserve le droit d'apporter des modifications à ce manuel et aux produits décrits à tout moment,
sans avertissement ni obligation. Les éditions révisées se trouvent sur le site Internet du fabricant.
Consignes de sécurité
A V I S
Le fabricant décline toute responsabilité quant aux dégâts liés à une application ou un usage inappropriés de ce
produit, y compris, sans toutefois s'y limiter, des dommages directs ou indirects, ainsi que des dommages
consécutifs, et rejette toute responsabilité quant à ces dommages dans la mesure où la loi applicable le permet.
L'utilisateur est seul responsable de la vérification des risques d'application critiques et de la mise en place de
mécanismes de protection des processus en cas de défaillance de l'équipement.
Veuillez lire l'ensemble du manuel avant le déballage, la configuration ou la mise en fonctionnement
de cet appareil. Respectez toutes les déclarations de prudence et d'attention. Le non-respect de
cette procédure peut conduire à des blessures graves de l'opérateur ou à des dégâts sur le matériel.
Assurez-vous que la protection fournie avec cet appareil n'est pas défaillante. N'utilisez ni n'installez
cet appareil d'une façon différente de celle décrite dans ce manuel.
Interprétation des indications de risques
D A N G E R
Indique une situation de danger potentiel ou imminent qui, si elle n'est pas évitée, entraîne des blessures graves,
voire mortelles.
A V E R T I S S E M E N T
Indique une situation de danger potentiel ou imminent qui, si elle n'est pas évitée, peut entraîner des blessures
graves, voire mortelles.
A T T E N T I O N
Indique une situation de danger potentiel qui peut entraîner des blessures mineures ou légères.
A V I S
Indique une situation qui, si elle n'est pas évitée, peut occasionner l'endommagement du matériel. Informations
nécessitant une attention particulière.
Etiquettes de mise en garde
Lisez toutes les informations et toutes les étiquettes apposées sur l’appareil. Des personnes peuvent
se blesser et le matériel peut être endommagé si ces instructions ne sont pas respectées. Un
30
Français
symbole sur l'appareil est référencé dans le manuel et accompagné d'une déclaration de mise en
garde.
Ceci est le symbole d'alerte de sécurité. Se conformer à tous les messages de sécurité qui suivent ce
symbole afin d'éviter tout risque de blessure. S'ils sont apposés sur l'appareil, se référer au manuel
d'utilisation pour connaître le fonctionnement ou les informations de sécurité.
Ce symbole indique un danger de pincement potentiel.
Le matériel électrique portant ce symbole ne doit pas être mis au rebut dans les réseaux domestiques
ou publics européens. Retournez le matériel usé ou en fin de vie au fabricant pour une mise au rebut
sans frais pour l'utilisateur.
Certification
Règlement canadien sur les équipements causant des interférences radio, IECS-003, Classe
A:
Les données d'essai correspondantes sont conservées chez le constructeur.
Cet appareil numérique de classe A respecte toutes les exigences du Règlement sur le matériel
brouilleur du Canada.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC part 15, limites de classe A :
Les données d'essai correspondantes sont conservées chez le constructeur. L'appareil est conforme
à la partie 15 de la règlementation FCC. Le fonctionnement est soumis aux conditions suivantes :
1. Cet équipement ne peut pas causer d'interférence nuisible.
2. Cet équipement doit accepter toutes les interférences reçues, y compris celles qui pourraient
entraîner un fonctionnement inattendu.
Les modifications de cet équipement qui n’ont pas été expressément approuvées par le responsable
de la conformité aux limites pourraient annuler l’autorité dont l’utilisateur dispose pour utiliser cet
équipement. Cet équipement a été testé et déclaré conforme aux limites définies pour les appareils
numériques de classe A, conformément à la section 15 de la réglementation FCC. Ces limites ont
pour but de fournir une protection raisonnable contre les interférences néfastes lorsque l’équipement
fonctionne dans un environnement commercial. Cet équipement génère, utilise et peut irradier
l'énergie des fréquences radio et, s'il n'est pas installé ou utilisé conformément au mode d'emploi, il
peut entraîner des interférences dangereuses pour les communications radio. Le fonctionnement de
cet équipement dans une zone résidentielle risque de causer des interférences nuisibles, dans ce
cas l'utilisateur doit corriger les interférences à ses frais Les techniques ci-dessous peuvent
permettre de réduire les problèmes d'interférences :
1. Débrancher l'équipement de la prise de courant pour vérifier s'il est ou non la source des
perturbations
2. Si l'équipement est branché sur le même circuit de prises que l'appareil qui subit des
interférences, branchez l'équipement sur un circuit différent.
3. Éloigner l'équipement du dispositif qui reçoit l'interférence.
4. Repositionner l’antenne de réception du périphérique qui reçoit les interférences.
5. Essayer plusieurs des techniques ci-dessus à la fois.
Français
31
Présentation du produit
D A N G E R
Dangers chimiques ou biologiques. Si cet instrument est utilisé pour la surveillance d'un procédé de
traitement et/ou d'un système de dosage de réactifs chimiques auxquels s'appliquent des limites
réglementaires et des normes de surveillance motivées par des préoccupations de santé et de sécurité
publiques ou de fabrication et de transformation d'aliments ou de boissons, il est de la responsabilité
de l'utilisateur de cet instrument qu'il connaisse et applique les normes en vigueur et qu'il ait à sa
disposition suffisamment de mécanismes pour s'assurer du bon respect de ces normes dans
l'éventualité d'un dysfonctionnement de l'appareil.
A T T E N T I O N
Risque d’incendie. Ce produit n'est pas adapté à l'utilisation avec des liquides inflammables.
L'échantillonneur portable SD900 collecte des échantillons liquides à intervalles réguliers et conserve
les échantillons dans des bouteilles ou des récipients. Utilisez l'échantillonneur pour un large éventail
d'applications aqueuses, pour des substances toxiques et des solides en suspension. Configurez
l'échantillonneur avec différentes bouteilles ou différents arrêtoirs ou récipients. Voir la Figure 1.
Figure 1 Présentation du produit
1 Base compacte 6 Connexions du transmetteur
2 Base standard isolée 7 Pompe
3 Pièce centrale 8 Contrôleur
4 Source d'alimentation 9 Capteur de liquide
5 Couvercle supérieure 10 Base standard isolée pour un récipient de 21 l
(5,5 gal)
Composants du produit
Assurez-vous d'avoir bien reçu tous les composants. Voir la Figure 2. Si des éléments manquent ou
sont endommagés, contactez immédiatement le fabricant ou un représentant commercial.
32
Français
Figure 2 Composants du produit
1 Base (standard, compacte ou composite) 6 Chargeur de batterie (en option)
2 Composants pour une option à bouteille unique (la
bouteille et le support peuvent changer)
7 Alimentation secteur (en option)
3 Pièce centrale avec transmetteur 8 Crépine
4 Batterie (en option) 9 Tube d'admission, en vinyle ou doublé en téflon
5 Couvercle supérieure 10 Composants pour une option à plusieurs bouteilles
(les bouteilles et les arrêtoirs peuvent changer)
Installation
D A N G E R
Dangers multiples. Seul le personnel qualifié doit effectuer les tâches détaillées dans cette section du
document.
Précautions concernant l'espace confiné
D A N G E R
Risque d’explosion Une formation portant sur les tests de pré-entrée, la ventilation, les procédures
d'entrée, les procédures d'évacuation/de sauvetage et les mesures de sécurité est nécessaire avant
d'entrer dans des lieux confinés.
Les informations suivantes sont fournies dans le but d'aider les utilisateurs à appréhender les
dangers et les risques associés aux espaces confinés.
Le 15 avril 1993, le règlement final de l'OSHA concernant le CFR 1910.146, Permit Required
Confined Spaces (Espaces confinés nécessitant l'autorisation), est devenue une loi. Cette norme
affecte directement plus de 250 000 sites industriels aux Etats-Unis et a été rédigée dans le but de
protéger la santé et la sécurité des travailleurs en espace confiné.
Définition d'un espace confiné :
Tout endroit ou clôture qui présente (ou est susceptible de présenter) une ou plusieurs des
conditions suivantes :
Une atmosphère qui contient une concentration d'oxygène inférieure à 19,5 % ou supérieure à
23,5 % et/ou une concentration de sulfure d'hydrogène (H
2
S) supérieure à 10 ppm.
Une atmosphère qui peut être inflammable ou explosive en présence de gaz, vapeurs, brumes,
poussières ou fibres.
Français
33
Des matériaux toxiques qui, en cas de contact ou d'inhalation, sont susceptibles d'occasionner
des blessures, des problèmes de santé ou la mort.
Les espaces confinés ne sont pas conçus pour l'occupation humaine. Les espaces confinés
disposent d’un accès limité et présentent des risques connus ou potentiels. Les trous d’homme, les
colonnes, les tuyaux, les cuves, les chambres de commutation et autres emplacements similaires
sont des exemples d’espaces confinés.
Il convient de toujours suivre les procédures de sécurité standard avant d'entrer dans des espaces
et/ou des endroits confinés soumis à des gaz dangereux, des vapeurs, des brumes, des poussières
ou des fibres Avant de pénétrer dans un espace confiné, veuillez lire l'ensemble des procédures
liées à l'accès.
Installation mécanique
Conseils d'installation sur site
D A N G E R
Risque d’explosion L'instrument n'est pas homologué pour une installation dans des zones
dangereuses.
Reportez-vous aux conseils ci-dessous pour l'évaluation de l'emplacement du site.
Suivez toutes les précautions de sécurité si l'échantillonneur est installé dans un espace confiné.
Voir la Précautions concernant l'espace confiné à la page 33.
Vérifiez que la température de l'emplacement ne dépasse pas la plage spécifiée. Voir la
Caractéristiques à la page 28.
Placez l'échantillonneur sur une surface plane ou suspendez l'échantillonneur à l'aide des sangles
de suspension, du palier de support ou de la clef tricoise. Voir la Installez l'échantillonneur dans le
trou d'homme à la page 34 et la documentation applicable pour l'installation.
Aussi près que possible de la source d'échantillon pour réduire le décalage d'analyse. Voir la
Branchement de l'échantillonneur à la page 37.
Pour connaître les limitations de vitesse durant le transport et la levée verticale maximale, voir la
Caractéristiques à la page 28.
Installez l'échantillonneur dans le trou d'homme
Installezl'échantillonneurau-dessus de l'eau de l'échantillon dans le trou d'homme.
Installezl'échantillonneuravec une clef tricoise ou un palier de support. Installez la clef tricoise dans le
trou d'homme. La clef tricoise est maintenue par pression contre le mur. Le palier de support
comporte la même largeur que le couvercle de trou d'homme. Installez le palier de support
directement sous le couvercle pour le maintenir.Consultez le manuel complet sur le site Web du
fabricant.Consultez la documentation fournie avec les accessoires pour installer l'échantillonneur.
Installez le distributeur ou la fermeture de bouteille pleine (en option)
L'ensemble du distributeur ou de la fermeture de bouteille pleine est généralement monté en usine.
Consultez la documentation du distributeur ou de la fermeture de bouteille pleine pour l'installation.
Préparation de l'échantillonneur
Nettoyez les bouteilles de l'échantillon
Nettoyez les capuchons et bouteilles d'échantillon avec une brosse, de l'eau et un détergent doux.
Rincez les conteneurs à l'eau douce, puis à l'eau distillée.
Installation d'une bouteille
Utilisez une seule bouteille pour prélever un échantillon composite. Lorsque la bouteille est pleine, le
système de fermeture de bouteille pleine interrompt le programme d'échantillonnage.
1. Nettoyez les bouteilles de l'échantillon. Voir Nettoyez les bouteilles de l'échantillon à la page 34.
2. Installez la bouteille d'échantillon tel qu'illustré sur la Figure 3.
34
Français
Figure 3 Installation de bouteille unique
1 Bouteille en polyéthylène, 10 l (2,5 gal) (1918) 6 Bouteille en polyéthylène, 21 l (5,5 gal) (6494)
2 Bouteille en verre, 10 l (2,5 gal) (6559) 7 Base compacte (8975)
3 Service après-vente (1502) 8 Base standard isolée (8976)
4 Bouteille en polyéthylène, 15 l (4 gal) (1367) 9 Base standard isolée pour une bouteille de 21 l
(5,5 gal) (8561)
5 Bouteille en polyéthylène, 19 l (5 gal) (6498)
Installez la première bouteille pour collecter des échantillons
Utilisez des bouteilles multiples pour collecter des échantillons dans des bouteilles séparées ou dans
plusieurs bouteilles. Le distributeur déplace le tube de prélèvement au-dessus de chaque bouteille.
Positionnez les bouteilles dans l'embase de l'échantillonneur comme indiqué sur la Figure 4. Installez
la première bouteille de l'échantillon (numéro 1) sous l'étiquette dans l'embase de l'échantillonneur.
Installez les bouteilles restantes en nombre croissant dans le sens indiqué par l'étiquette. Voir sur la
Figure 5 à la page 37 le schéma des composants requis.
Français
35
Figure 4 Installation de la bouteille numéro 1
1 Emplacement de la bouteille numéro 1 (base
compacte)
6 Sangles élastiques
2 Emplacement de la bouteille numéro 1 pour
24 bouteilles
7 Base standard isolée pour une bouteille de 21 l
(5,5 gal)
3 Emplacement de la bouteille numéro 1 pour
8 bouteilles
8 Base standard isolée
4 Emplacement de la bouteille numéro 1 pour 2 ou
4 bouteilles
9 Base compacte
5 Arrêtoir
Installation de plusieurs bouteilles
Lorsque plusieurs bouteilles sont installées, un bras de distribution déplace le tube d'échantillon au-
dessus de chaque bouteille. Le prélèvement d'échantillons s'interrompt automatiquement lorsque le
nombre spécifié d'échantillons a été prélevé.
1. Nettoyez les bouteilles de l'échantillon. Voir Nettoyez les bouteilles de l'échantillon à la page 34.
2. Installez la première bouteille de l'échantillon (numéro 1) sous l'étiquette dans l'embase de
l'échantillonneur. Voir Installez la première bouteille pour collecter des échantillons à la page 35.
3. Assemblez les bouteilles d'échantillon tel qu'illustré sur la Figure 5. Pour huit bouteilles ou plus,
assurez-vous que la première bouteille se place à côté de l'indicateur de la première bouteille
dans le sens des aiguilles d'une montre.
4. Placez l'assemblage de bouteilles dans l'échantillonneur. Pour huit bouteilles ou plus, alignez les
câbles dans les emplacements du plateau inférieur.
36
Français
Figure 5 Installation de bouteilles multiples
1 Positionneur/arrêtoir (2347) 8 Bouteilles en verre, 3,8 l (1 gal), (4x) (2216) ou
bouteilles en polyéthylène, 3,8 l (1 gal), (4x) (2217)
2 Arrêtoir (1422) 9 Bouteilles en verre, 1,9 l (0,5 gal), (8x) (1118) ou
bouteilles en polyéthylène, 2,3 l (0,61 gal), (8x)
(657)
3 Arrêtoir (2190) 10 Bouteilles en polyéthylène, 1 l (0,26 gal), (24x)
(737)
4 Arrêtoir (2189) 11 Bouteilles en verre, 350 ml (11,83 oz) (24x) (732)
5 Bouteilles en verre, 950 ml (32,12 oz) (8x) (2348) 12 Base compacte (8975)
6 Bouteilles en polyéthylène, 575 ml (19,44 oz), (24x)
(1369)
13 Base standard (8976)
7 Bouteilles en verre, 3,8 l (1 gal), (2x) (2214) ou
bouteilles en polyéthylène, 3,8 l (1 gal), (2x) (2215)
Branchement de l'échantillonneur
Monter la crépine au milieu du jet de l'échantillonneur (pas près de la surface ou près du fond) pour
s'assurer qu'un échantillon représentatif est collecté. Voir la Figure 6. Voir la Figure 7pour installer le
tube d'admission.
1. Connectez le tube à l'échantillonneur tel qu'illustré sur la Figure 7.
Remarque : Utilisez le kit de connexion 2186 si un tube en Téflon est utilisé.
2. Mettez le tube d'admission et la crépine dans le flux principal de la source de l'échantillon où l'eau
est turbulente et bien mélangée.
Raccourcissez autant que possible le tube d'admission.
Placez le tube d'admission verticalement afin que le tube se vide entièrement entre les
échantillons.
Remarque : si une pente verticale est impossible ou si le tube est pressurisé, désactivez le capteur de
liquide. Etalonnage manuel du volume de l'échantillon. Voir la Etalonnage manuel du volume de l'échantillon
à la page 48.
Français
37
Assurez-vous que le tube d'admission n'est pas pincé.
Figure 6 Installation de l'instrument
1 Crépine 3 Tube d'admission
2 Relevage vertical
Figure 7 Installation du tube d'admission
Installation électrique
Connexions du transmetteur
A V E R T I S S E M E N T
Risque d'électrocution. Tout équipement externe relié doit avoir fait l'objet d'un contrôle de sécurité
conformément aux normes nationales applicables.
La Figure 8 illustre les connexions pouvant être effectuées avec le transmetteur.
38
Français
Figure 8 Connecteurs du transmetteur
1 Appareil auxiliaire 4 Appareil SDI-12 en option
2 Alimentation 5 Distributeur/fermeture de bouteille pleine
3 Communications série
Branchement d'un débitmètre
Branchez un débitmètre au transmetteur pour démarrer ou interrompre l'échantillonneur lorsque le
flux d'échantillon est supérieur ou inférieur à la valeur spécifiée.
Articles à réunir :
Câble universel complet pour les débitmètres Sigma (ou demi-câble 980 pour le débitmètre de
modèle 980).
Répartiteur en option pour les branchements additionnels. Deux répartiteurs ou plus peuvent être
connectés en série.
1. Connectez une extrémité du câble audébitmètre. Pour le modèle de débitmètre 980, reportez-
vous au manuel de l'utilisateur.
2. Branchez l'autre extrémité du câble au connecteur d'appareil auxiliaire du transmetteur.
Remarque : si le débitmètre dispose d'une câble à 6 broches, utilisez le câble d'adaptateur 6 à 7 broches.
Connexion d'un PC ou d'un réseau de communication
Connectez un PC ou un réseau de communication au transmetteur pour transférer des données ou
configurer l'échantillonneur.
Articles à réunir :
Câble série, RS232 vers DB-9 7 broches (voir Figure 9). Connexions : B vers 5 (masse du signal) ;
D vers 3 (RDC) ; F vers 2 (TXD) ; G (masse).
1. Reliez une extrémité du câble série au réseau ou à l'appareil de communication.
2. Branchez l'autre extrémité du câble au connecteur de communication série du transmetteur.
Figure 9 Câble de communication série
Français 39
Branchement de l'échantillonneur à l'alimentation
D A N G E R
Risque d'électrocution Si cet équipement est utilisé à l'extérieur ou dans des lieux potentiellement
humides, un disjoncteur de fuite à la terre (GFCI/GFI) doit être utilisé pour le branchement de
l'équipement à sa source d'alimentation secteur.
D A N G E R
Risque d’incendie. Installez un disjoncteur 15 A sur la ligne électrique. Le disjoncteur peut servir à
couper l'alimentation localement s'il est situé tout près de l'équipement.
D A N G E R
Risque d'électrocution Un raccordement à la terre est nécessaire.
A V E R T I S S E M E N T
Risque d'électrocution Assurez-vous de disposer d'un accès facile à la coupure d'alimentation locale.
A V E R T I S S E M E N T
Risque de décharge électrique. L'alimentation peut surchauffer quand le délai entre les cycles de
pompe est trop court.
A V I S
Assurez-vous que la durée du cycle de la pompe ne provoquera pas la surchauffe de l'alimentation électrique.
Voir la Calcul de la durée des cycles de la pompe à la page 41.
Connectez l'échantillonneur à une batterie ou à une alimentation secteur. Reportez-vous aux
illustrations suivantes.
40
Français
Calcul de la durée des cycles de la pompe
La pompe doit rester arrêtée pendant une durée minimum entre les cycles d'échantillonnage, afin
que l'alimentation électrique ne surchauffe pas.
1. Déterminez la durée totale pendant laquelle la pompe fonctionnera de façon continue au cours
d'un cycle d'échantillonnage. Inclure toutes les étapes : pré-purge, rinçage de l'admission,
échantillonnage, échantillonnages répétés et post-purge.
2. Déterminez la durée minimale pendant laquelle la pompe doit être arrêtée entre les cycles
d'échantillonnage à l'aide de la Figure 10.
3. Assurez-vous que le rythme ou l'intervalle du programme de l'échantillonneur permet l'arrêt de la
pompe pendant la durée minimale déterminée entre les cycles d'échantillonnage.
Figure 10 Durées d'arrêt/fonctionnement de la pompe à 50 °C (122 °F)
1 Durée (en minutes) pendant laquelle la pompe doit
rester arrêtée
3 Exemple : si la pompe fonctionne de façon continue
pendant 5 minutes, la pompe doit rester arrêtée
pendant 15 minutes.
2 Durée (en minutes) pendant laquelle la pompe
fonctionne de façon continue
Mise sous tension
Appuyez sur le bouton d'alimentation pour mettre l'appareil sous ou hors tension. Assurez-vous que
l'alimentation (alimentation CA ou alimentation par batterie) est correctement installée.
Interface utilisateur et navigation
Le clavier du transmetteur est illustré à la Figure 11. Utilisez la touche MENU pour configurer les
programmes d'échantillonnage et les paramètres du transmetteur ou réaliser les tests de diagnostic.
Utilisez les flèches ainsi que les touches ENTER (Entrée) et BACK (Retour) pour parcourir le menu,
sélectionner des éléments et saisir des valeurs. Vérifiez les flèches en bas ou à côté de l'écran pour
savoir si d'autres options sont disponibles.
Français
41
Figure 11 Clavier du transmetteur
1 MARCHE 5 MENU 9 STATUT
2 VOLUME CALIBRATION
(Etalonnage du volume)
6 ENTER (Entrée) 10 RUN/HALT PROGRAM
(Lancer/Interrompre le
programme)
3 STOP (Arrêt) 7 BACK (Retour) 11 MANUAL OPERATION
(Fonctionnement manuel)
4 Diode électroluminescente 8 FLECHES
Fonctionnement
Configuration des paramètres système
1. Appuyez sur MENU.
2. Accédez à SYSTEM SETUP (Configuration système).
3. Modifiez les options applicables.
Option Description
Réglage heure/date Permet de définir l'heure (format 24 h) et la date.
Communication Permet de sélectionner la vitesse en baud (19 200, 38 400, 57 600 ou 115 200)
et le protocole (Modbus RTU ou ASCII) pour le port série.
Configuration de la base Permet de sélectionner la base d'échantillonneur (réfrigéré tout climat, réfrigéré,
portable standard ou portable compact).
Langue Permet de sélectionner la langue du transmetteur.
Et capteur liq Permet d'étalonner le capteur liquide ou de sélectionner les valeurs par défaut.
Régler le contraste Permet de régler le contraste de l'écran LCD.
Cycle de fonctionnement
de la pompe
Permet de modifier le cycle de fonctionnement de la pompe. Plage : de 50 à
100 % (100 % par défaut).
42 Français
Option Description
Configuration du mot de
passe
Permet de configurer un mot de passe pour contrôler l'accès aux menus de
configuration du programme et de configuration système. Sélectionnez
ENABLE (Activer) > YES (Oui) pour définir un nouveau mot de passe ou
ENABLE (Activer) > NO (Non) pour activer le mot de passe existant. Si le mot
de passe est défini pour la première fois, utilisez 900900 en tant que mot de
passe actuel.
Durée de vie de la
tuyauterie
Permet de définir un rappel lorsque la tuyauterie de la pompe doit être
remplacée. Accédez à ENABLE (Activer) > RESET CYCLES (Réinitialiser
cycles) pour réinitialiser le nombre de cycles de la pompe sur 0. Accédez à
ENABLE (Activer) > CYCLE LIMIT (Limite de cycle) pour modifier le nombre de
cycles de la pompe utilisé pour le rappel. Accéder à ENABLE (Activer) >
STATUS (Etat) pour afficher le nombre actuel de cycles de la pompe.
SDI-12 (en option) Permet de configurer le fonctionnement d'un appareil SDI-12. Cette option
s'affiche uniquement lorsqu'un appareil est détecté.
Configuration d'un programme d'échantillonnage
Un programme d'échantillonnage comprend tous les paramètres nécessaires pour prélever
automatiquement des échantillons. Prélevez les échantillons à intervalles réguliers ou lorsque le
programme d'échantillonnage est terminé.
1. Appuyez sur MENU.
2. Accédez à PROGRAM SETUP (Configuration du programme) > MODIFY ALL (Modifier tout). Le
premier écran s'affiche.
3. Sélectionnez ou saisissez les valeurs pour chaque paramètre. Utilisez les touches fléchées
gauche et droite pour déplacer le curseur à gauche ou à droite. Utilisez les touches fléchées haut
et bas pour changer la valeur. Changez les unités au besoin (gal ou ml) avec les touches
fléchées.
Option Description
Bouteilles Quantité de bouteilles : le nombre de bouteilles dans l'échantillonneur (1, 2, 4, 8 ou
24).
Fermeture de bouteille pleine : activé/désactivé. Si activé, le transmetteur vérifie les
conditions de débordement de la bouteille.
Volume de la bouteille : la capacité de chaque bouteille. Plage : 50 à 65 000 ml
(0,01 à 17,17 gal)
Tuyauterie
d'admission
Longueur de la tuyauterie : la longueur du tube d'admission depuis la crépine
jusqu'au capteur de liquide. Plage : 100 à 3 000 cm ou 3 à 99 pieds. Une longueur
précise est nécessaire pour obtenir un volume d'échantillon précis.
Type de tube d'admission : le diamètre et le matériau du tube d'admission (¼ po.
vinyle, 3/8 po. vinyle ou 3/8 po. Teflon).
Retard de
programme
Activer/désactiver : une fois activé, le programme d'échantillonnage démarre à une
heure ou un nombre de comptages spécifié. Date et heure : la date et l'heure de
démarrage du programme (format 24 h). Comptages : le nombre de comptages ou
de pulsations d'un débitmètre lorsque le programme démarre (1 à 9 999 comptages).
Si le rythme de l'échantillon est défini ultérieurement sur l'heure, le retard de
programme sera désactivé.
Français 43
Option Description
Rythme/collecte
d'échantillon
Basés sur le temps—Reportez-vous à Collecte basée sur le temps à la page 44.
Remarque : Lorsque CTVV est utilisé :
Le capteur de liquide est toujours activé.
La distribution d'échantillon est composite.
Le mode d'exécution est non continu.
Le premier échantillon est pris après l'intervalle.
La source du débit est toujours l'entrée 4–20 mA
Basés sur le débit—Reportez-vous à Collecte basée sur le débit à la page 45.
Distribution
d'échantillon
Fournir des échantillons à tous > YES (Oui)
Chaque échantillon sera livré dans toutes les bouteilles. Terminer après le dernier :
le programme s'interrompt après le dernier échantillon (1 à 999). Continu : le
programme continue jusqu'à ce qu'il soit interrompu manuellement.
Fournir des échantillons à tous > NO (Non)
Les échantillons sont livrés à un sous-réseau de bouteilles . Echantillons/bouteille :
nombre d'échantillons à prélever dans chaque bouteille (1 à 999).
Bouteilles/échantillon : nombre de bouteilles qui contiendront le même échantillon.
Capteur de liquide Activé ou désactivé. Si activé, le volume d'échantillon doit être calibré selon le
temps.
Sample volume
(Volume
d’échantillon)
Le volume de chaque échantillon en ml (1 gal = 3 785,4 ml). Si le mode
bouteilles/échantillon est sélectionné, chaque bouteille reçoit un volume entier
d'échantillon (10 à 10 000 ml). Assurez-vous que le volume d'échantillon ne dépasse
pas le volume de bouteille. Les volumes d'échantillon sont arrondis aux 10 ml les
plus proches.
Rinçages
d'admission
Le nombre de rinçages du tube d'admission avant prélèvement d'un échantillon (0 à
3).
Tentatives
d'échantillonnage
Le nombre d'essais d'échantillonnage après échec d'un échantillonnage (0 à 3).
ID site Le nom d'un emplacement d'échantillonnage (jusqu'à 12 caractères). L'ID du site est
utilisé comme le nom prédéfini si le programme est enregistré.
Échantillonnage
avancé
Permet d'accéder au menu d'échantillonnage avancé. Si nécessaire, configurez les
options dans le menu d'échantillonnage avancé.
Lancement du
programme
La création du programme d'échantillonnage basique se termine. L'échantillonneur
invite l'utilisateur à démarrer ou annuler le programme.
Collecte basée sur le temps
1. Sélectionnez ou saisissez les valeurs pour chaque option.
Option Description
Intervalle de rythme Prélève les échantillons à intervalles réguliers, en heures et minutes
(0:01–999:00).
44 Français
Option Description
Prendre le premier
échantillon
Le programme démarre immédiatement ou après le premier intervalle.
Volume variable Indique si les volumes d'échantillon varient (en fonction du débit). Si oui,
reportez-vous à Utiliser le volume variable à la page 45. Si non, reportez-vous
à Ne pas utiliser de volume variable à la page 45.
Utiliser le volume variable
Le rythme d'échantillonnage est volume variable à intervalle fixe (CTVV).
1. Sélectionnez ou saisissez les valeurs pour chaque option.
Option Descriptions
Sélectionner l'unité Sélectionnez les unités de débit utilisées.
Définir l'entrée 4 mA Saisissez le débit qui correspond à l'entrée 4 mA
Définir l'entrée 20 mA Saisissez le débit qui correspond à l'entrée 20 mA
Débit moyen Indiquez le débit moyen pour le site (1–999 999). Les unités sont indiquées
par les unités sélectionnées ci-dessus.
Volume désiré pour la durée Indiquez le volume total devant être collecté lors du programme
d'échantillonnage (10–10 000 ml).
Période de collecte Indiquez la période de collecte totale en heures et minutes (0:01–999:00).
Volume minimum
d'échantillon
Si le volume de débit n'est pas suffisant pour collecter ce volume minimum,
l'échantillon est ignoré (10–10 000 ml).
Ne pas utiliser de volume variable
1. Sélectionnez ou saisissez les valeurs pour l'option.
Option Description
Prendre le premier échantillon Vous permet de sélectionner si le programme commence immédiatement
ou à la fin du premier intervalle.
Collecte basée sur le débit
Spécifiez la source du débit pour l'échantillonnage basé sur le débit :
Comptages—Reportez-vous à Utiliser les comptages pour la source de débit à la page 45.
Entrée 4–20 mA—Reportez-vous à Utilisez l'entrée de 4–20 mA pour la source de débit
à la page 46.
Utiliser les comptages pour la source de débit
1. Sélectionnez ou saisissez les valeurs pour chaque option.
Option Description
Prendre l'échantillon chaque Prélève des échantillons à intervalles de débit réguliers, en comptages (1 à
9 999 comptages).
Durée de temporisation Prélève un échantillon si le volume de débit est très faible.
Français 45
Option Description
Durée Durée maximale entre les échantillons (0:01–999:00). Le retardateur
démarre à nouveau après chaque prélèvement d'échantillon.
Prendre le premier
échantillon
Le programme démarre immédiatement ou après le premier intervalle de
temps.
Utilisez l'entrée de 4–20 mA pour la source de débit
1. Sélectionnez ou saisissez les valeurs pour chaque option.
Option Descriptions
Sélectionner l'unité Permet de sélectionner les unités de débit utilisées.
Définir l'entrée 4 mA Saisissez le débit qui correspondrait à l'entrée 4 mA.
Définir l'entrée 20 mA Saisissez le débit qui correspondrait à l'entrée 20 mA.
Prendre l'échantillon
chaque
Saisissez le débit total après lequel l'échantillon doit être tiré.
Temps de dépassement Sélectionnez Activé pour forcer la collecte d'un échantillon si le volume de débit
est exceptionnellement bas.
Durée Durée maximale entre les échantillons (0:01–999:00). Le retardateur démarre à
nouveau après chaque prélèvement d'échantillon.
Volume d’échantillon Entrez le volume à collecter par échantillon. Si le mode bouteilles/échantillon
est sélectionné, chaque bouteille reçoit un volume entier d'échantillon (100 à
10 000 ml). Cette option est ignorée si le volume variable est sélectionné.
Restauration des paramètres par défaut
L'option de restauration définit tous les paramètres du programme sur les paramètres par défaut.
1. Appuyez sur MENU.
2. Accédez à PROGRAM SETUP (Configuration du programme) > RESTORE (Restaurer).
3. Sélectionnez YES (Oui).
Lancer ou arrêter un programme
Utilisez la touche RUN/HALT PROGRAM (Lancer/Interrompre le programme) pour lancer ou
interrompre un programme. Jusqu'à 12 heures/dates de démarrage/arrêt peuvent être définis par
l'utilisateur. Un programme doit être arrêté avant que la configuration du programme ou du système
puisse être modifiée.
1. Appuyez sur RUN/HALT PROGRAM (Lancer/Interrompre le programme).
2. Sélectionnez l’une des options.
Option Description
Début Démarrez le programme acutellement chargé. L'état change en RUNNING
(En service).
Halt (Interrompre) Interrompt le programme temporairement. L'état change en HALTED
(Interrompu).
Reprendre : Le programme reprend à partir de l'endroit auquel il a été interrompu. L'état
change en RUNNING (En service).
Start over (Reprendre) Le programme reprend depuis le début. L'état change en RUNNING (En
service).
End program (Terminer le
programme)
Termine le programme en cours. L'état change en COMPLETE (Terminé).
46 Français
Fonctionnement manuel
Utilisez le fonctionnement manuel pour collecter un échantillon instantané, déplacer le bras du
distributeur ou actionner la pompe. Pour plus d'informations, reportez-vous à la version enrichie de
ce manuel.
Calibrage de volume
Le volume d'échantillon peut être étalonné automatiquement avec le capteur de liquide ou
manuellement. Si le capteur de liquide est utilisé pour le calibrage de volume, le capteur est ajusté
pour mesurer avec précision tous les volumes programmés. Si le capteur de liquide est désactivé, le
volume pour les échantillons, les rinçages et l'eau de précipitation doit être étalonné manuellement.
Pour des applications typiques, l'eau du robinet peut être utilisée pour l'étalonnage de volume. Si la
composition d'un échantillon est sensiblement différente des échantillons d'eau typiques, étalonnez
l'échantillonneur en utilisant le liquide à prélever. Tous les programmes doivent être interrompus
avant l'étalonnage.
Vous permet de calibrer le capteur de liquide
La précision du capteur de liquide peut changer en fonction du type de liquide en cours
d'échantillonnage. Par exemple, le volume d'un échantillon fortement turbide peut ne pas être aussi
précis que celui d'une eau claire. Le capteur de liquide peut être étalonné avec l'eau d'échantillon
afin d'améliorer la précision du volume.
1. Placez le tube d'admission dans l'eau d'échantillon.
2. Appuyez sur MENU.
3. Accédez à PROGRAM SETUP (Configuration du programme) > MODIFY SELECTED (Modifier la
sélection) > LIQUID SENSOR (Capteur de liquide) > ENABLE (Activer) pour vous assurer que le
capteur de liquide est activé.
4. Accédez à SYSTEM SETUP (Configuration du système) > LIQ SENSOR CAL. (Etal. du capteur
de liquide).
5. Sélectionnez PERFORM CAL. (Effectuer étal.).
6. Sélectionnez START (Démarrer). La pompe fonctionne en sens inverse pour purger le tube
d'admission. Elle fonctionne alors en avant.
7. Vérifiez le flux d'échantillon dans le tube d'admission afin de vous assurer que l'échantillon
traverse le capteur de liquide.
8. Lorsque l'échantillon traverse le capteur de liquide et qu'il n'y a aucune bulle, appuyez sur la
touche STOP (Arrêter). Le tube d'admission est automatiquement purgé. L'étalonnage du capteur
de liquide est terminé.
9. Sélectionnez DONE (Terminé) pour quitter ou REPEAT (Répéter) pour effectuer à nouveau
l'étalonnage.
Restauration de l'étalonnage d'usine
L'étalonnage de volume pour le capteur de liquide peut être réinitialisé aux paramètres par défaut.
1. Sélectionnez PROGRAM SETUP (Configuration du programme) > MODIFY SELECTED (Modifier
la sélection) > LIQUID SENSOR (Capteur de liquide) > ENABLE (Activer) pour vous assurer que
le capteur de liquide est activé.
2. Appuyez sur VOLUME CALIBRATION (Etalonnage du volume) et sélectionnez RESET CAL.
(Réinitialiser Etal.). L'étalonnage du volume est défini sur les paramètres par défaut. Le message
« value has been reset to 0 » (valeur réinitialisée sur 0) s'affiche.
Etalonnage du volume d'échantillon avec le capteur de liquide
Etalonnez le volume d'échantillon avec le capteur de liquide activé pour ajuster les volumes
d'échantillon avec précision.
Français
47
1. Accédez à PROGRAM SETUP (Configuration du programme) > MODIFY SELECTED (Modifier la
sélection) > LIQUID SENSOR (Capteur de liquide) > ENABLE (Activer) pour vous assurer que le
capteur de liquide est activé.
2. Placez le tube d'admission dans l'eau d'échantillon ou de l'eau du robinet.
Remarque : utilisez l'eau d'échantillon pour une plus grande précision.
3. Placez le tuyau d'évacuation de la pompe dans un cylindre gradué.
4. Appuyez sur VOLUME CALIBRATION (Etalonnage du volume) et sélectionnez CALIBRATION
(Etalonnage).
5. Sélectionnez START (Démarrer). La pompe fonctionne en sens inverse pour purger le tube
d'admission. Elle fonctionne alors en avant et collecte le volume d'échantillon spécifié dans le
programme d'échantillonnage. La pompe fonctionne en sens inverse pour purger le tube
d'admission.
6. Lorsque l'échantillon a été collecté, sélectionnez DONE (Terminé). Comparez le volume
d'échantillon dans le cylindre gradué avec le volume d'échantillon saisi dans le menu de
configuration du programme.
7. Si le volume qui a été collecté est différent du volume d'échantillon dans le programme en cours,
entrez le volume qui a été réellement collecté. Appuyez sur ENTER. Le capteur est ajusté pour
mesurer avec précision tous les volumes programmés.
8. Sélectionnez START (Démarrer) pour lancer le programme d'échantillonnage ou CANCEL
(Annuler) pour quitter le menu d'étalonnage du volume.
Etalonnage manuel du volume de l'échantillon
Lorsque le capteur de liquide est désactivé, tous les volumes doivent être étalonnés manuellement
en fonction de la durée. Le volume d'échantillon est calibré pour le volume indiqué dans le
programme en cours. Si le volume d'échantillon est modifié dans le programme, le volume
d'échantillon doit être à nouveau étalonné pour le nouveau volume.
1. Sélectionnez PROGRAM SETUP (Configuration du programme) > MODIFY SELECTED (Modifier
la sélection) > LIQUID SENSOR (Capteur de liquide) > DISABLE (Désactiver) pour vous assurer
que le capteur de liquide est désactivé.
2. Placez le tube d'admission dans l'eau d'échantillon ou de l'eau du robinet.
3. Placez le tuyau d'évacuation de la pompe dans un cylindre gradué.
4. Appuyez sur VOLUME CALIBRATION (Etalonnage du volume). Si plusieurs volumes sont
affichés, sélectionnez l'un d'eux pour l'étalonnage.
5. Sélectionnez START (Démarrer). La pompe fonctionne en sens inverse pour purger le tube
d'admission. Elle fonctionne alors en avant et lance le prélèvement de l'échantillon.
6. Arrêtez la pompe au volume choisi ou rincez l'endroit :
Volume d'échantillon : appuyez sur STOP (Arrêter) lorsque le volume spécifié dans le
programme principal est prélevé.
Rincer : appuyez sur STOP (Arrêter) lorsque le liquide dans le tube d'admission atteint le
détecteur de liquide.
Volume d'eaux de précipitation : appuyez sur STOP (Arrêter) lorsque le volume spécifié dans
le programme d'eaux de précipitation est prélevé.
7. Sélectionnez DONE (Terminé) pour accéder au prochain étalonnage de volume ou REPEAT
(Répéter) pour effectuer à nouveau l'étalonnage.
8. Lorsque tous les volumes ont été étalonnés, sélectionnez DONE (Terminé).
9. Reliez le tube d'évacuation au raccord de tube sur l'échantillonneur.
10. Sélectionnez START (Démarrer) pour lancer le programme d'échantillonnage ou CANCEL
(Annuler) pour quitter.
48
Français
Vérification du volume d'échantillon
Pour vous assurer que le volume d'échantillon est précis, prenez un échantillon instantané. Ne
revenez pas dans le calibrage pour vérifier le volume, puisque la compensation de volume est
remise à zéro au début d'un calibrage.
1. Appuyez sur MANUAL OPERATION (Fonctionnement manuel).
2. Sélectionnez GRAB SAMPLE (Echantillon instantané.
3. Placez le tube d'admission dans l'eau d'échantillon.
4. Placez le tuyau d'évacuation de la pompe dans un cylindre gradué.
5. Si le capteur de liquide est activé, saisissez le volume à vérifier.
6. Si le capteur de liquide est désactivé, entrez le volume indiqué dans le programme.
7. Appuyez sur ENTER. Le cycle de pompe démarre (purge-échantillon-purge).
8. Comparez le volume qui a été prélevé dans le cylindre gradué au volume attendu. Si le volume
prélevé est incorrect, effectuez à nouveau l'étalonnage.
Maintenance
D A N G E R
Dangers multiples. Seul le personnel qualifié doit effectuer les tâches détaillées dans cette section du
document.
D A N G E R
Risque d'électrocution Coupez l'alimentation de l'instrument avant d'effectuer des activités de
maintenance ou d'entretien.
A V E R T I S S E M E N T
Exposition aux risques biologiques. Respectez des protocoles de manipulation sécurisés lorsque vous
êtes en contact avec les flacons d'échantillon et les composants de l'échantillonneur.
A V E R T I S S E M E N T
Dangers multiples. Le technicien doit s'assurer que l'équipement fonctionne correctement et de
manière sécurisée après avoir exécuté une procédure de maintenance.
Nettoyage de l'instrument
A T T E N T I O N
Risque d’incendie. Ne pas utiliser d'agents inflammables pour nettoyer l'appareil.
A V I S
Ne nettoyez pas le radiateur du compartiment du transmetteur avec des liquides, toutes sortes confondues.
Si l'eau n'est pas suffisante pour nettoyer le transmetteur et la pompe, déconnectez le transmetteur
et éloignez-le de l'échantillonneur. Laissez sécher le transmetteur et la pompe avant de réinstaller les
pièces et de les remettre en service.
Nettoyez l'échantillonneur comme suit :
Boîtier de l'échantillonneur : nettoyez les surfaces internes et externes avec un tissu humide et un
détergent doux. N'utilisez pas de nettoyeurs abrasifs ou de dissolvants.
Français
49
Bouteilles d'échantillon : nettoyez la bouteille avec une brosse, de l'eau et un détergent doux.
Rincez avec de l'eau douce, puis avec de l'eau distillée. Utilisez un autoclave pour nettoyer les
bouteilles en verre si nécessaire.
Remplacement du dessiccateur
Une cartouche déshydratante dans le transmetteur absorbe l'humidité et empêche la corrosion.
Surveillez la couleur du déshydratant à travers la fenêtre (Figure 12). Un déshydratant non utilisé est
jaune. Lorsque la couleur devient verte, remplacez le déshydradant.
1. Desserrez et retirez la cartouche déshydratante (Figure 12).
2. Retirez le bouchon et le déshydratant utilisé.
3. Remplissez le tube du dessiccant avec un dessiccant frais.
4. Placez le bouchon.
5. Mettez de la graisse sur le joint torique.
6. Installez le tube du déshydratant dans le transmetteur.
Figure 12 Cartouche desséchante
1 Prise 3 Joint torique
2 Ensemble du 4 Fenêtre du dessiccant
Maintenance de la pompe
A T T E N T I O N
Risque de pincement. Coupez l'alimentation de l'instrument avant d'effectuer des activités de
maintenance ou d'entretien.
Remplacement du tuyau de la pompe
A V I S
L'utilisation d'une tuyauterie autre que celle fournie par le fabricant peut causer une usure excessive des pièces
mécaniques et/ou des performance moindres de la pompe.
Recherchez sur la tuyauterie de la pompe des traces d'usure au niveau de la zone de friction des
rouleaux avec la tuyauterie. Remplacez la tuyauterie lorsque le message de rappel s'affiche ou
lorsque la tuyauterie montre des signes d'usure.
Prérequis :
Tuyauterie de la pompe – prédécoupée ou complète, 4,6 m ou 15,2 m (15 ou 50 pieds)
1. Déconnectez l'alimentation du transmetteur.
2. Si la tuyauterie complète est utilisée, coupez-la et ajoutez des points d'alignement. Consultez le
manuel complet sur le site Web du fabricant.
3. Retirez la tuyauterie de la pompe tel qu'illustré dans les étapes suivantes.
50
Français
4. Nettoyez les résidus de silicone à l'intérieur du corps de la pompe et sur les rouleaux.
5. Installez la nouvelle tuyauterie de la pompe tel qu'illustré dans les étapes suivantes.
Français 51
Nettoyage du rotor
Nettoyez le rotor, les voies de tube de pompe et le boîtier de pompe avec un détergent doux.
Reportez-vous au chapitre Remplacement du tuyau de la pompe à la page 50 et aux étapes
illustrées suivantes.
Remplacement du tube du bras de distribution
Le bras du distributeur se déplace au-dessus de chaque bouteille pendant l'échantillonnage des
bouteilles multiples. Remplacez le tube dans le bras de distribution lorsque le tube est usé. Assurez-
vous que le tube correct est utilisé pour le distributeur ainsi que le bras de distributeur adéquats.
Remarque : Le tube du distributeur n'est pas le même que celui de la pompe. Si vous installez le tube de la pompe
dans le distributeur, ce dernier peut être endommagé. Les échantillons ne pourront pas être correctement prélevés
car le bras du distributeur ne se déplacera pas aisément.
1. Retirez le tube du bras de distribution et du haut de la pièce centrale.
2. Insérez le nouveau tube dans le bras de distribution. Assurez-vous que le tube est aligné selon
les dimensions spécifiées dansFigure 13.
3. Insérez l'autre extrémité du tube dans le raccord sur le haut de la pièce centrale.
4. Effectuez le test de diagnostic pour le distributeur afin de vous assurer que le système fonctionne
correctement.
Figure 13 Ensemble distributeur
1 Buse 3 Moteur de distributeur 5 Tube distributeur
2 Bras de distributeur 4 Axe
52 Français
Remplacement des fusibles
L'utilisateur ne peut pas remplacer les fusibles. Contactez l’assistance technique. Un fusible grillé
peut indiquer que l'échantillonneurest défaillant et qu'une réparation est nécessaire.
Français 53
Tabla de contenidos
Especificaciones en la página 54 Funcionamiento en la página 68
Información general en la página 56 Mantenimiento en la página 75
Instalación en la página 59
Información adicional
En el sitio web del fabricante encontrará información adicional.
Especificaciones
Las especificaciones están sujetas a cambios sin previo aviso.
Muestreador portátil SD900
Especificación Base estándar Base compacta Base compuesta
Dimensiones Diámetro: 50,5 cm (19,9 pulg.)
Altura: 69,4 cm (27,3 pulg.)
Diámetro: 44,1 cm
(17,4 pulg.)
Altura: 61 cm (24 pulg.)
Diámetro: 50,28 cm
(19,8 pulg.)
Altura: 79,75 cm
(31,4 pulg.)
Peso 15 kg (35,6 lb) con botellas de
polietileno de 1 l (24x)
14,8 kg (32,6 lb) con recipiente de
polietileno de 10 l (2,5 gal) (1x)
12,2 kg (27 lb) con botellas
de polietileno de 575 ml
(19,44 oz) (24x)
12,9 kg (28,3 lb) con
recipiente de polietileno de
10 l (2,5 gal) (1x)
15 kg (36 lb) con
botellas de vidrio de
950 ml (32,12 oz) (12x)
Carcasa ABS resistente a impactos, estructura de 3 piezas; base con pared doble con aislamiento
de 2,54 cm (1 pulg.) Aislamiento: contacto directo de la botella con hielo.
Temperatura de la
muestra
0-60 °C (32-140 °F)
Filtros De acero inoxidable 316 en tamaño estándar, de alta velocidad o bajo perfil para
aplicaciones a poca profundidad y de Teflon
®
o acero inoxidable 316 en tamaño estándar.
Tubos de entrada
de muestras
D.I. de 9,5 mm (3/8 pulg.), polietileno con revestimiento de vinilo o teflón
Capacidad de la
botella para
muestras
Botellas de polietileno de 1 l
(0,26 gal) y/o botellas de vidrio de
350 ml (11,83 oz) (24x)
Botellas de polietileno de 2,3 l
(0,6 gal) y/o botellas de vidrio de
1,9 l (0,5 gal) (8x)
Botellas de polietileno de 3,8 l
(1 gal) y/o botellas de vidrio de
3,8 l (1 gal) (4x)
Botellas de polietileno de 3,8 l
(1 gal) y/o botellas de vidrio de
3,8 l (1 gal) (2x)
Recipiente de polietileno
compuesto de 21 l (5,5 gal) o 15-L
(4 gal), de polietileno de 20 l
(5,25 gal), de polietileno de 10 l
(2,5 gal) o de vidrio de 10 l
(2,5 gal) (1x)
Botellas de polietileno de
575 ml (19,44 oz) (24x)
Botellas de vidrio de 950 ml
(32,12 oz) (8x)
Botella de polietileno de
10 l (2,5 gal) (1x)
Botella de vidrio de 10 l
(2,5 gal) (1x)
Botella de polietileno de
21 l (5,5 gal) (1x)
54 Español
Controlador SD
Especificación Detalles
Dimensiones (an. x alt. x prof.) 29,2 x 17,1 x 26,4 cm (11½ x 6¾ x 10
3
/
8
pulg.)
Peso 4.2 kg (9.26 libras)
Carcasa Mezcla de PC/ABS, NEMA 4X, 6, IP 67
Requisitos de alimentación 15 V CC suministrados por una fuente de alimentación 8754500;
15 V CC suministrados por una fuente de alimentación integrada
Protección contra sobrecarga 7 A, fusible de la línea CC para la bomba
Temperatura de funcionamiento 0 a 50 °C (32 a 122 °F)
Temperatura de almacenamiento –30 a 60 °C (–22 a 140 °F)
Humedad en
almacenamiento/funcionamiento
Condensación 100%
Bomba Peristáltica de alta velocidad, con rodillos de nilatron montados
sobre resortes
Recinto de la bomba IP37
Tubería de la bomba 9,5 mm de diámetro interior x 15,9 mm de diámetro exterior (
3
/
8
pulg. de diámetro interior x
5
/
8
pulg. de diámetro exterior) de
silicona
Vida de la tubería de la bomba Ciclos de 20.000 muestras con: volumen de muestras de 1 l
(0,3 galones), 1 enjuague, intervalo de cadenciado de 6 minutos,
4,9 m (16 pies) del tubo de entrada de
3
/
8
pulg. , 4,6 m (15 pies)
de elevación vertical, temperatura de la muestra de 21 °C (70 °F)
Elevación de la muestra vertical 8,5 m (28 pies) máximo para: 8,8 m (29 pies) del tubo de entrada
de vinilo de
3
/
8
pulg. a nivel del mar a 20–25 °C (68–77 °F)
Caudal de la bomba 4,8 l/min (1,25 gpm) a 1 m (3 pies) de elevación vertical con un
tubo de entrada de
3
/
8
pulg. típico
Volumen de la muestra Se puede programa en incrementos de 10 ml (0,34 onzas) desde
10 hasta 10.000 ml (3,38 onzas a 2,6 galones)
Repetibilidad del volumen de la muestra
(típica)
±5% del volumen de la muestra de 200 ml usando detección de
líquido sin calibrar con: elevación vertical de 4,6 m (15 pies),
4,9 m (16 pies) de un tubo de entrada de vinilo de
3
/
8
pulg. , una
sola botella, dispositivo de corte para botella llena a temperatura
ambiente y elevación de 1524 m (5000 pies)
Precisión del volumen de la muestra (típico) ±10% del volumen de la muestra de 200 ml usando detección de
líquido sin calibrar con: elevación vertical de 4,6 m (15 pies),
4,9 m (16 pies) de un tubo de entrada de vinilo de
3
/
8
pulg. , una
sola botella, dispositivo de corte para botella llena a temperatura
ambiente y elevación de 1524 m (5000 pies)
Modos de muestreo Intervalos: de tiempo fijo, de caudal fijo, de tiempo variable, de
caudal variable y de caudal variable en tiempo constante (CTVV,
por sus siglas en inglés).
Consulte la información sobre muestras de caudal variable en
tiempo constante (CTVV) en la versión ampliada de este manual.
Distribución: compuesta en una sola botella, compuesta en varias
botellas, aislada en varias botellas, botellas por muestra,
muestras por botella, combinación de botellas por muestra,
muestras por botella
Modos de ejecución Continuos o no continuos, con cantidad de muestras ingresada
por el usuario
Español 55
Especificación Detalles
Múltiples programas Almacena hasta tres programas de muestreo.
Velocidad de transferencia (típica) 0,9 m/s (2,9 pies/s) con: elevación vertical de 4,6 m (15 pies),
4,9 m (16 pies) de tubos de entrada de vinilo de
3
/
8
pulg. , 21 °C
(70 °F) y elevación de 1524 m (5000 pies)
Sensor de líquido Ultrasónico. Cuerpo: Ultem
®
aprobado según norma NSF ANSI
51, cumple con USP Clase VI
Histórico de muestreo Hasta 510 registros
Purga de aire Aire purgado automáticamente antes y después de cada muestra.
La duración compensa automáticamente las variaciones en las
longitudes del tubo de entrada.
Registro de eventos 510 registros
Conexiones Alimentación, auxiliar, comunicaciones serie, distribuidor, SDI-12
Materiales impregnados Acero inoxidable, polietileno, teflón, Ultem, silicona
Comunicaciones RS232, Modbus, SDI-12
Garantía 1 año
Información general
En ningún caso el fabricante será responsable de ningún daño directo, indirecto, especial, accidental
o resultante de un defecto u omisión en este manual. El fabricante se reserva el derecho a modificar
este manual y los productos que describen en cualquier momento, sin aviso ni obligación. Las
ediciones revisadas se encuentran en la página web del fabricante.
Información de seguridad
A V I S O
El fabricante no es responsable de ningún daño debo a un mal uso de este producto incluyendo, sin limitación,
daños directos, fortuitos o circunstanciales y reclamaciones sobre los daños que no estén recogidos en la
legislación vigente. El usuario es el responsable de la identificación de los riesgos críticos y de tener los
mecanismos adecuados de protección de los procesos en caso de un posible mal funcionamiento del equipo.
Lea todo el manual antes de desembalar, instalar o trabajar con este equipo. Ponga atención a
todas las advertencias y avisos de peligro. El no hacerlo puede provocar heridas graves al usuario o
daños al equipo.
Asegúrese de que la protección proporcionada por el equipo no está dañada. No utilice ni instale
este equipo de manera distinta a lo especificado en este manual.
Uso de la información sobre riesgos
P E L I G R O
Indica una situación potencial o de riesgo inminente que, de no evitarse, provocará la muerte o lesiones graves.
A D V E R T E N C I A
Indica una situación potencial o inminentemente peligrosa que, de no evitarse, podría provocar la muerte o
lesiones graves.
P R E C A U C I Ó N
Indica una situación potencialmente peligrosa que podría provocar una lesión menor o moderada.
A V I S O
Indica una situación que, si no se evita, puede provocar daños en el instrumento. Información que requiere
especial énfasis.
56 Español
Etiquetas de precaución
Lea todas las etiquetas y rótulos adheridos al instrumento. En caso contrario, podrían producirse
heridas personales o daños en el instrumento. El símbolo que aparezca en el instrumento se
comentará en el manual con una declaración de precaución.
Este es un símbolo de alerta de seguridad. Obedezca todos los mensajes de seguridad que se
muestran junto con este símbolo para evitar posibles lesiones. Si se encuentran sobre el instrumento,
consulte el manual de instrucciones para obtener información de funcionamiento o seguridad.
Este símbolo indica un peligro de pellizco potencial.
En Europa, el equipo eléctrico marcado con este símbolo no se debe desechar mediante el servicio
de recogida de basura doméstica o pública. Devuelva los equipos viejos o que hayan alcanzado el
término de su vida útil al fabricante para su eliminación sin cargo para el usuario.
Certificación
Reglamentación canadiense sobre equipos que provocan interferencia, IECS-003, Clase A
Registros de pruebas de control del fabricante.
Este aparato digital de clase A cumple con todos los requerimientos de las reglamentaciones
canadienses para equipos que producen interferencias.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC Parte 15, Límites Clase "A"
Registros de pruebas de control del fabricante. Este dispositivo cumple con la Parte 15 de las
normas de la FCC estadounidense. Su operación está sujeta a las siguientes dos condiciones:
1. El equipo no puede causar interferencias perjudiciales.
2. Este equipo debe aceptar cualquier interferencia recibida, incluyendo las interferencias que
pueden causar un funcionamiento no deseado.
Los cambios o modificaciones a este equipo que no hayan sido aprobados por la parte responsable
podrían anular el permiso del usuario para operar el equipo. Este equipo ha sido probado y
encontrado que cumple con los límites para un dispositivo digital Clase A, de acuerdo con la Parte
15 de las Reglas FCC. Estos límites están diseñados para proporcionar una protección razonable
contra las interferencias perjudiciales cuando el equipo está operando en un entorno comercial. Este
equipo genera, utiliza y puede irradiar energía de radio frecuencia, y si no es instalado y utilizado de
acuerdo con el manual de instrucciones, puede causar una interferencia dañina a las radio
comunicaciones. La operación de este equipo en un área residencial es probable que produzca
interferencia dañina, en cuyo caso el usuario será requerido para corregir la interferencia bajo su
propio cargo. Pueden utilizarse las siguientes técnicas para reducir los problemas de interferencia:
1. Desconecte el equipo de su fuente de alimentación para verificar si éste es o no la fuente de la
interferencia.
2. Si el equipo está conectado a la misma toma eléctrica que el dispositivo que experimenta la
interferencia, conecte el equipo a otra toma eléctrica.
3. Aleje el equipo del dispositivo que está recibiendo la interferencia.
4. Cambie la posición de la antena del dispositivo que recibe la interferencia.
5. Trate combinaciones de las opciones descritas.
Español
57
Descripción general del producto
P E L I G R O
Peligro químico o biológico. Si este instrumento se usa para controlar un proceso de tratamiento y/o un
sistema de suministro químico para el que existan límites normativos y requisitos de control
relacionados con la salud pública, la seguridad pública, la fabricación o procesamiento de alimentos o
bebidas, es responsabilidad del usuario de este instrumento conocer y cumplir toda normativa
aplicable y disponer de mecanismos adecuados y suficientes que satisfagan las normativas vigentes
en caso de mal funcionamiento del equipo.
P R E C A U C I Ó N
Peligro de incendio. Este producto no ha so diseñado para utilizarse con líquidos inflamables.
El muestreador portátil SD900 recoge muestras líquidas a intervalos específicos y conserva las
muestras en botellas o recipientes. Utilice el muestreador para una amplia variedad de aplicaciones
acuosas, así como para recoger contaminantes tóxicos y sólidos en suspensión. Configure el
muestreador con diferentes dispositivos de retención, botellas o recipientes. Consulte Figura 1.
Figura 1 Descripción general del producto
1 Base compacta 6 Conexiones del controlador
2 Base aislada estándar 7 Bomba
3 Sección central 8 Controlador
4 Fuente de alimentación 9 Sensor de líquido
5 Tapa superior 10 Base aislada estándar para recipiente de 21 l
(5,5 gal)
Componentes del producto
Asegúrese de haber recibido todos los componentes. Consulte la Figura 2. Si faltan artículos o están
dañados, póngase en contacto con el fabricante o el representante de ventas inmediatamente.
58
Español
Figura 2 Componentes del producto
1 Base (estándar, compacta o compuesta) 6 Cargador de batería (opcional)
2 Componentes para la opción de botella única (la
botella y el soporte se pueden cambiar)
7 Fuente de alimentación de CA (opcional)
3 Sección central con controlador 8 Filtro
4 Batería (opcional) 9 Tubo de entrada, revestimiento de vinilo o teflón
5 Tapa superior 10 Componentes para la opción de varias botellas
(las botellas y los dispositivos de retención se
pueden cambiar)
Instalación
P E L I G R O
Peligros diversos. Sólo el personal cualificado debe realizar las tareas descritas en esta sección del
documento.
Precauciones para espacios confinados
P E L I G R O
Peligro de explosión. XXX
La información que se incluye a continuación se ofrece para ayudar a los usuarios a comprender los
peligros y riesgos asociados a los espacios confinados.
El 15 de abril de 1993, el dictamen definitivo de la OSHA (Administración de Seguridad y Salud
Ocupacional) sobre los Espacios Confinados que Requieren Permiso para Ingresar (CFR 1910.146),
se hizo ley. Esta nueva norma afecta directamente a más de 250.000 sitios industriales de los
Estados Unidos, y fue creada con el fin de proteger la salud y la seguridad de los trabajadores en
espacios confinados.
Definición de espacio confinado:
Un espacio confinado es cualquier lugar o recinto que presente (o tenga la posibilidad inmediata de
presentar) una o más de las siguientes condiciones:
Una atmósfera con una concentración de oxígeno que sea inferior al 19,5% o superior al 23,5%
y/o una concentración de sulfuro de hidrógeno (H
2
S) superior a 10 ppm.
Español
59
Una atmósfera que pueda ser inflamable o explosiva debido a gases, vapores, nieblas, polvos o
fibras.
Materiales tóxicos que, ante el contacto o la inhalación, puedan provocar lesiones, el deterioro de
la salud o la muerte.
Los espacios confinados no están destinados a ser ocupados por seres humanos. Los espacios
confinados tienen entrada restringida y contienen riesgos conocidos o potenciales. Como ejemplos
de espacios confinados encontramos las bocas de inspección, las chimeneas, los caños, las tinas,
los armarios de distribución y demás lugares similares.
Antes de entrar en espacios confinados y/o lugares con presencia de gases, vapores, nieblas,
polvos o fibras peligrosos, se deben seguir siempre procedimientos de seguridad estándares. Antes
de entrar en un espacio confinado, lea todos los procedimientos relacionados con la entrada a
espacios confinados.
Instalación mecánica
Pautas para la instalación
P E L I G R O
Peligro de explosión. El instrumento no está aprobado para su instalación en lugares peligrosos.
Consulte en las instrucciones siguientes el modo de evaluar la ubicación.
Siga todas las precauciones de seguridad si el muestreador está instalado en un espacio
reducido. Consulte Precauciones para espacios confinados en la página 59.
Asegúrese de que la temperatura del emplazamiento está dentro del rango de especificaciones.
Consulte Especificaciones en la página 54.
Instale el muestreador en una superficie nivelada o cuelgue el muestreador con el arnés de
suspensión, el soporte de arqueta o la barra de extensión. Consulte Instalación el muestreador en
una boca de inspección en la página 60 y la documentación de instalación pertinente.
Lo más cerca posible de la fuente de muestras para reducir el retraso en el análisis. Consulte
Conexión del muestreador en la página 63.
Para obtener información sobre la velocidad del transporte y la inclinación vertical máxima,
consulte Especificaciones en la página 54.
Instalación el muestreador en una boca de inspección
Instale el muestreador por encima de la muestra de agua en una boca de inspección. Instale el
muestreador con una barra de extensión o un soporte de arqueta. Instale la barra de extensión en la
boca de inspección. La barra de extensión se sujeta por la presión que se ejerce contra las paredes.
El soporte de arqueta tiene la misma anchura que la cubierta de la boca de inspección. Instale el
soporte de arqueta directamente por debajo de la cubierta a modo de apoyo.Consulte el manual
completo en la página web del fabricante.Consulte la documentación suministrada con los
accesorios para instalar el muestreador.
Instalación del distribuidor o dispositivo de corte por botella llena (opcional)
El conjunto del distribuidor o dispositivo de corte por botella llena generalmente viene instalado de
fábrica. Consulte la documentación del distribuidor o del dispositivo de corte por botella llena para
obtener información sobre la instalación.
Preparación del tomamuestras
Limpieza de las botellas para muestras
Limpie las botellas para muestras y los tapones con un cepillo, agua y detergente suave. Enjuague
los recipientes con agua corriente y luego con agua destilada.
60
Español
Instalación de una única botella
Utilice una única botella para recoger la muestra compuesta. Cuando la botella está llena, el
dispositivo de corte por botella llena detiene el programa de muestreo.
1. Limpie las botellas para muestras. Consulte la Limpieza de las botellas para muestras
en la página 60.
2. Instale la botella para muestras como se indica en la Figura 3.
Figura 3 Instalación de una única botella
1 Botella de polietileno, 10 l (2,5 gal) (1918) 6 Botella de polietileno, 21 l (5,5 gal) (6494)
2 Botella de vidrio, 10 l (2,5 gal) (6559) 7 Base compacta (8975)
3 Asistencia técnica (1502) 8 Base aislada estándar (8976)
4 Botella de polietileno, 15 l (4 gal) (1367) 9 Base aislada estándar para botella de 21 l (5,5 gal)
(8561)
5 Botella de polietileno, 19 l (5 gal) (6498)
Instalación de la primera botella para recogidas de muestras múltiples
Utilice múltiples botellas para recoger muestras en botellas distintas o en más de una botella. El
distribuidor mueve el tubo para muestras por encima de cada botella. Instale las botellas en la base
del muestreador, tal como se muestra en la Figura 4. Instale la primera botella para muestras
(número 1) por debajo de la etiqueta que está en la base del muestreador. Instale el resto de las
botellas siguiendo una numeración creciente, en la dirección indicada en la etiqueta. Consulte la
Figura 5 en la página 63 para obtener un diagrama de los componentes requeridos.
Español
61
Figura 4 Instalación de la botella número 1
1 Ubicación de la botella número 1 (base compacta) 6 Bandas elásticas
2 Ubicación de la botella número 1 para 24 botellas 7 Base aislada estándar para botella de 21 l (5,5 gal)
3 Ubicación de la botella número 1 para 8 botellas 8 Base aislada estándar
4 Ubicación de la botella número 1 para 2 o 4 botellas 9 Base compacta
5 Dispositivo de retención
Instalación de varias botellas
Cuando se instalan varias botellas, un brazo del distribuidor mueve el tubo para muestras por
encima de cada botella. La recogida de muestras se detiene automáticamente cuando ya se ha
recogido el número especificado de muestras.
1. Limpie las botellas para muestras. Consulte la Limpieza de las botellas para muestras
en la página 60.
2. Instale la primera botella para muestras (número 1) por debajo de la etiqueta que está en la base
del muestreador. Consulte Instalación de la primera botella para recogidas de muestras múltiples
en la página 61.
3. Coloque las botellas para muestras como se indica en la Figura 5. Si hay ocho botellas o más,
compruebe que la primera botella está junto al indicador de primera botella y que se colocan en
el sentido de las agujas del reloj.
4. Coloque el conjunto de botellas en el tomamuestras. Si hay ocho botellas o más, alinee los
cables en las ranuras de la bandeja inferior.
62
Español
Figura 5 Instalación para múltiples botellas
1 Regulador de posición/dispositivo de retención
(2347)
8 Botellas de vidrio, 3,8 l (1 gal) (4x) (2216) o botellas
de polietileno, 3,8 l (1 gal) (4x) (2217)
2 Dispositivo de retención (1422) 9 Botellas de vidrio, 1,9 l (0,5 gal) (8x) (1118) o
botellas de polietileno, 2,3 l (0,61 gal) (8x) (657)
3 Dispositivo de retención (2190) 10 Botellas de polietileno, 1 l (0,26 gal) (24x) (737)
4 Dispositivo de retención (2189) 11 Botellas de vidrio, 350 ml (11,83 oz) (24x) (732)
5 Botellas de vidrio, 950 ml (32,12 oz) (8x) (2348) 12 Base compacta (8975)
6 Botellas de polietileno, 575 ml (19,44 oz) (24x)
(1369)
13 Base estándar (8976)
7 Botellas de vidrio, 3,8 l (1 gal) (2x) (2214) o botellas
de polietileno, 3,8 l (1 gal) (2x) (2215)
Conexión del muestreador
Instale el filtro en el medio de la corriente de la muestra (no cerca de la superficie o del fondo) para
asegurarse de que recoge una muestra representativa. Consulte Figura 6. Consulte la Figura 7 para
instalar el tubo de entrada.
1. Conecte los tubos al tomamuestras como se indica en la Figura 7.
Nota: Si usa tubos con revestimiento de teflón, emplee el kit de conexión 2186.
2. Coloque el tubo de entrada y el filtro en la corriente principal de la fuente de origen de la
muestra, donde el agua es turbulenta y está bien mezclada.
El tubo de entrada debe ser lo más corto posible.
Mantenga el tubo de entrada en pendiente vertical, para que se vacíe completamente entre
muestra y muestra.
Nota: Si no se puede colocar en pendiente vertical o si el tubo se encuentra bajo presión, desactive el
sensor de líquido. Calibre el volumen de la muestra de forma manual. Consulte Calibración del volumen de
la muestra de forma manual en la página 74.
Compruebe que el tubo de entrada no esté aplastado.
Español
63
Figura 6 Instalación del instrumento
1 Filtro 3 Tubo de entrada
2 Elevación vertical
Figura 7 Instalación del tubo de entrada
Instalación eléctrica
Conexiones del controlador
A D V E R T E N C I A
Riesgo de descargas eléctricas. El equipo conectado de forma externa debe someterse a una
evaluación estándar de seguridad aplicable.
En la Figura 8 se muestran las conexiones que se pueden realizar en el controlador.
64
Español
Figura 8 Conectores del controlador
1 Dispositivo auxiliar 4 Opción de dispositivo SDI-12
2 Fuente de alimentación 5 Dispositivo de corte para botella llena/distribuidor
3 Comunicaciones en serie
Conexión de un caudalímetro
Conecte un caudalímetro al controlador para iniciar o detener el tomamuestras cuando el flujo de la
muestra descienda por debajo de un valor especificado o lo supere.
Recopilación de elementos:
Cable completo multiuso para los caudalímetros Sigma (o cable semicompleto 980 para el
caudalímetro modelo 980).
Repartidor opcional para realizar conexiones adicionales. Se pueden conectar en serie dos o más
repartidores.
1. Conecte un extremo del cable al caudalímetro. En el caso del caudalímetro del modelo 980,
consulte el manual del usuario del caudalímetro del modelo 980.
2. Conecte el otro extremo del cable al conector del dispositivo auxiliar en el controlador.
Nota: Si el caudalímetro tiene un cable de 6 patillas, utilice el cable adaptador de 6 patillas a 7 patillas.
Conexión de un PC o una red de comunicaciones
Conecte un PC o una red de comunicaciones al controlador para transferir datos o configurar el
tomamuestras.
Recopilación de elementos:
Cable serie, RS232 de 7 patillas a DB-9 (consulte la Figura 9). Conexiones: B a 5 (puesta a tierra
de señal); D a 3 (RCD); F a 2 (TXD); G (puesta a tierra).
1. Conecte un extremo del cable serie al dispositivo o red de comunicaciones.
2. Conecte el otro extremo del cable al conector de comunicaciones serie del controlador.
Figura 9 Cable de comunicaciones serie
Español 65
Conexión el muestreador a la alimentación eléctrica
P E L I G R O
Peligro de electrocución. Si este equipo se usa en exteriores o en lugares potencialmente húmedos,
debe utilizarse un disyuntor de interrupción de circuito por falla a tierra (GFCI/GFI) para conectar el
equipo a la alimentación eléctrica.
P E L I G R O
Peligro de incendio. Instale un disyuntor de 15 A en la línea de alimentación. Un disyuntor puede ser la
desconexión de alimentación local, si se encuentra en las proximidades del equipo.
P E L I G R O
Peligro de electrocución. Se requiere una conexión de toma a tierra (PE).
A D V E R T E N C I A
Peligro de electrocución. Asegúrese de que sea fácil acceder a la desconexión de alimentación local.
A D V E R T E N C I A
Riesgo de electrocución. El suministro de alimentación se puede recalentar cuando el tiempo entre
ciclos de bombeo es muy corto.
A V I S O
Asegúrese de que la duración del ciclo de bombeo no provoca la sobrecarga de la fuente de alimentación.
Consulte Cálculo del tiempo de ciclo de la bomba en la página 67.
Conecte el muestreador a una batería o a una toma de alimentación de CA. Consulte los pasos
ilustrados que se muestran a continuación.
66
Español
Cálculo del tiempo de ciclo de la bomba
La bomba debe permanecer apagada durante un período mínimo de tiempo entre los ciclos de
muestreo, para que la fuente de alimentación no se sobrecaliente.
1. Calcule la cantidad total de tiempo que la bomba funcionará sin interrupción durante un ciclo de
muestreo. Incluya todas las etapas: purga previa, lavado de la línea de admisión, muestreo,
reintentos de muestreo y purga posterior.
2. Determine la cantidad mínima de tiempo que la bomba debe estar apagada entre los ciclos de
muestreo con ayuda de la Figura 10.
3. El cadenciado o el intervalo de tiempo del programa del tomamuestras debe hacer que la bomba
se apague entre los ciclos de muestreo durante el período de tiempo mínimo correspondiente.
Figura 10 Tiempos de encendido y apagado de la bomba a 50 °C (122 °F)
1 Tiempo (en minutos) durante los cuales la bomba
debe permanecer apagada
3 Ejemplo: si la bomba funciona sin interrupción
durante 5 minutos, debe permanecer apagada
15 minutos.
2 Tiempo (en minutos) durante los cuales la bomba
funciona sin interrupción
Conexión de la alimentación
Pulse el botón de encendido para encender o apagar el dispositivo. Compruebe que el suministro de
alimentación (alimentación de CA o pilas) se ha instalado correctamente.
Interfaz del usuario y navegación
El teclado del controlador se muestra en la Figura 11. Utilice la tecla MENU (Menú) para configurar
los programas de muestreo, definir los ajustes del controlador o llevar a cabo las pruebas de
diagnóstico. Utilice las teclas de flechas, ENTER (Intro) y BACK (Retroceso) para desplazarse por
los menús, realizar selecciones e introducir valores. Mire si aparecen flechas en la parte inferior o en
el lateral de la pantalla para saber si hay más opciones disponibles.
Español
67
Figura 11 Teclado del controlador
1 POWER (Alimentación) 5 MENU (Menú) 9 STATUS (Estado)
2 VOLUME CALIBRATION
(Calibración del volumen)
6 ENTER (Intro) 10 RUN/HALT PROGRAM
(Ejecutar/Detener programa)
3 STOP (Parada) 7 BACK (Atrás) 11 MANUAL OPERATION
(Funcionamiento manual)
4 LED 8 TECLAS DE FLECHA
Funcionamiento
Configuración de los ajustes del sistema
1. Pulse MENU (Menú).
2. Vaya a SYSTEM SETUP (Configuración del sistema).
3. Cambie las opciones correspondientes.
Opción Descripción
Time/date setup (Configurar
hora/fecha)
Ajuste la hora (formato de 24 horas) y la fecha.
Communication
(Comunicación)
Seleccione la velocidad en baudios (19200, 38400, 57600 o 115200) y el
protocolo (Modbus RTU o ASCII) para el puerto serie.
Setup base (Configurar base) Seleccione la base del muestreador (refrigerado para todo tipo de clima,
refrigerado, estándar portátil o compacto portátil).
Language (Idioma) Seleccione el idioma del controlador.
Liq sensor cal (Calibración del
sensor de líquido)]
Calibre el sensor de líquidos o ajústelo a la configuración predeterminada.
Set contrast (Ajustar
contraste)
Ajuste el contraste de la pantalla LCD.
Pump duty cycle (Ciclo de
trabajo de la bomba)
Cambie el ciclo de trabajo de la bomba. Rango: entre 50% y 100% (valor
predeterminado: 100%).
68 Español
Opción Descripción
Password setup (Configurar
contraseña)
Defina una contraseña para controlar el acceso a los menús de
configuración del sistema y del programa. Seleccione ENABLE>YES
(Activar>Sí) para establecer una nueva contraseña o ENABLE>NO
(Activar>No) para activar la contraseña ya existente. Si se define la
contraseña por primera vez, utilice 900900 como la contraseña actual.
Tubing life (Vida de la tubería) Defina un recordatorio que indique cuándo se deben cambiar los tubos de
la bomba. Desplácese a ENABLE>RESET CYCLES (Activar>Restablecer
ciclos) para poner a cero (0) el recuento de ciclos de la bomba. Vaya a
ENABLE>CYCLE LIMIT (Activar>Límite de ciclo) para cambiar el número
de ciclos de la bomba que debe emplear el recordatorio. Desplácese a
ENABLE>STATUS (Activar>Estado) para mostrar el número actual de
recuentos de ciclo de la bomba.
SDI–12 (optional)
[SDI–12 (opcional)]
Configure el funcionamiento de un dispositivo SDI–12. Esta opción se
muestra únicamente cuando se encuentra un dispositivo.
Configuración de un programa de muestreo
En un programa de muestreo se incluyen todos los parámetros que son necesarios para recoger
muestras de forma automática. Recoja las muestras a intervalos regulares o cuando se complete el
programa de muestreo.
1. Pulse MENU (Menú).
2. Desplácese a PROGRAM SETUP>MODIFY ALL (Configuración del programa>Modificar todo).
Aparecerá la primera pantalla.
3. Realice una selección o introduzca los valores para cada parámetro. Use las teclas de flecha
izquierda y derecha para desplazarse hacia la izquierda o hacia la derecha. Use las teclas de
flecha arriba y bajo para cambiar un valor. Si es necesario, cambie las unidades (galones o ml)
usando las teclas de flecha.
Opción Descripción
Bottles (Botellas) Cantidad de botellas: número de botellas que hay en el muestreador (1, 2, 4,
8 o 24).
Full BTL shutoff: Enabled/Disabled (Dispositivo de corte para botella llena:
activado/desactivado). Si está activado, el controlador comprueba si la
botella se desborda.
Bottle volume (Volumen de la botella): capacidad volumétrica de cada
botella. Rango: 50–65 000 ml (0,01 - 17,17 gal)
Intake tubing (Tubos de
entrada)
Tubing length (Longitud de los tubos): longitud de los tubos de entrada
desde el filtro hasta el sensor de líquido. Rango: 100–3000 cm p 3–99 pies.
Es necesario indicar una longitud exacta para poder obtener un volumen de
la muestra preciso.
Intake tube type (Tipo del tubo de entrada): diámetro y material de los tubos
de entrada (¼ pulg. de vinilo, 3/8 pulg. de vinilo o 3/8 pulg. de teflón).
Program delay (Retardo del
programa)
Enable/disable (Activar/desactivar): si está activado, el programa de
muestreo comienza en un momento específico o tras un número de
recuentos determinado. Date and time (fecha y hora): fecha y hora en la que
empieza el programa (formato de 24 horas). Counts (Recuentos): número de
recuentos o impulsos de un caudalímetro cuando el programa comienza (1 a
9999 recuentos). Si el cadenciado de muestreo se establece posteriormente
en base al tiempo, el retardo del programa se desactivará.
Español 69
Opción Descripción
Ritmo
muestra/colecc.muestra
Recogida de base tiempo: consulte Recogida de base tiempo
en la página 70.
Nota: Cuando se utiliza CTVV:
El sensor de líquido está siempre activado.
La distribución de la muestra es compuesta.
El modo de ejecución es discontinuo.
La primera muestra se toma tras el intervalo.
La fuente de flujo es siempre una entrada de 4-20 mA.
Por base flujo: consulte Recogida por base flujo en la página 71.
Sample distribution
(Distribución de la muestra)
Deliver samples to all>YES (Entregar muestras a todos>Sí)
Se entrega cada muestra a todas las botellas. End after last (Terminar tras
última): el programa se detiene después de la última muestra (1–999).
Continuous (Continuo): el programa continúa hasta que se detiene de forma
manual.
Deliver samples to all>NO (Entregar muestras a todos>No)
Las muestras se entregan a un subconjunto de botellas . Samples/bottle
(Muestras/botellas): número de muestras que hay que recoger en cada
botella (1-999). Bottles/sample (Botellas/muestras): número de botellas que
contendrán la misma muestra.
Liquid sensor (Sensor de
líquido)
Enable (Activado) o disabled (desactivado). Si está deshabilitado, el volumen
de la muestra debe ser calibrado por tiempo.
Sample volume (Volumen
de la muestra)
El volumen de cada muestra en ml (1 galón = 3785,4 ml). Si se está
utilizando el modo botellas/muestra, cada botella recibe un volumen
completo de la muestra (10-10.000 ml). Asegúrese de que el volumen de la
muestra no exceda el volumen de la botella. Los volúmenes de la muestra
se redondean a los 10 ml más cercanos.
Intake rinses (Lavados del
tubo de entrada)
El número de veces que el tubo de entrada se lava antes de recoger una
muestra (0–3).
Sample retries (Reintentos
de muestreo)
El número de veces que se intenta un muestreo antes de que se produzca
un fallo de muestreo (0–3).
Site ID (ID del sitio) El nombre de la ubicación de muestreo (máximo 12 caracteres). Si se
guarda el programa, la ID del sitio se usa como el nombre preestablecido
para el mismo.
muestreo avanzado Va al menú de muestreo avanzado. Si es necesario, configure las opciones
en el menú Advanced Sampling (Muestreo avanzado).
Ejec. programa (Run
program)
Finaliza la creación del programa de muestreo básico. El tomamuestras pide
al usuario que inicie o cancele el programa.
Recogida de base tiempo
1. Seleccione o introduzca los valores para cada opción.
Opción Descripción
Intervalo Recoja las muestras a intervalos de tiempo regulares, en ho ras y minutos
(0:01–999:00).
70 Español
Opción Descripción
Tome 1a muestra El programa se inicia de inmediato o tras el primer intervalo.
Volumen variable Especifique si varían los volúmenes de muestras (por caudal). En caso afirmativo,
consulte Uso de un volumen variable en la página 71. En caso negativo, consulte Sin
uso de un volumen variable en la página 71.
Uso de un volumen variable
El ritmo de muestra se expresa en tiempo constante volumen variable (CTVV).
1. Seleccione o introduzca los valores para cada opción.
Opción Descripción
Selecc. unidad Seleccione las unidades de caudal que desea utilizar.
Entr. 4mA MAP Introduzca la velocidad de caudal correspondiente para la entrada de 4 mA.
Entr. 20mA MAP Introduzca la velocidad de caudal correspondiente para la entrada de 20 mA.
Caudal promedio Especifique la velocidad de caudal media de la instalación (1–999 999). Las
unidades se especifican en función de las unidades seleccionadas arriba.
Time volume desired
(volumen y tiempo
deseado)
Especifique el volumen total que debe recogerse durante el programa de
muestreo (10–10 000 ml).
Collection period (Período
de recogida)
Especifique el período de recogida total en horas y minutos (0:01–999:00).
Vol min muestreo Si el volumen del caudal no es suficiente para recoger este volumen mínimo,
se omitirá la muestra (10–10 000 ml).
Sin uso de un volumen variable
1. Seleccione o introduzca los valores para la opción.
Opción Descripción
Tome 1a muestra Seleccione si desea que el programa se inicie de inmediato o tras el primer intervalo de
tiempo.
Recogida por base flujo
Especifique la fuente de flujo para la toma de muestras de base flujo:
Conteo: consulte Uso de recuentos para determinar la fuente de flujo en la página 71.
Entr. 4-20 mA: consulte Uso de una entrada de 4-20 mA para la fuente de flujo en la página 72.
Uso de recuentos para determinar la fuente de flujo
1. Seleccione o introduzca los valores para cada opción.
Opción Descripción
Muestra cada Recoja muestras a intervalos de caudal regulares, en recuentos (recuentos de 1–9999).
Tiempo anulac. Recoja una muestra si el volumen de caudal es muy bajo.
Español 71
Opción Descripción
Tiempo Tiempo máximo entre muestras (0:01–999:00). El temporizador vuelve a comenzar tras
haber recogido cada muestra.
Tome 1a muestra El programa se inicia de inmediato o tras el primer intervalo de tiempo.
Uso de una entrada de 4-20 mA para la fuente de flujo
1. Seleccione o introduzca los valores para cada opción.
Opción Descripción
Selecc. unidad Seleccione las unidades de caudal que desea utilizar.
Entr. 4mA MAP Introduzca la velocidad de caudal correspondiente para la entrada de 4 mA.
Entr. 20mA MAP Introduzca la velocidad de caudal correspondiente para la entrada de 20 mA.
Muestra cada Introduzca el caudal total a partir del cual se obtendrá la muestra.
Tiempo anulac. Seleccione Habilitar para recoger una muestra si el volumen del caudal es
anormalmente bajo.
Tiempo Tiempo máximo entre muestras (0:01–999:00). El temporizador vuelve a comenzar tras
haber recogido cada muestra.
Volumen muestra Ingresar el volumen a recoger por muestra. Si está seleccionado el modo
Botella/muestra, cada botella recibe un volumen completo de la muestra
(100-10 000 ml). Esta opción se omite si se selecciona un volumen variable.
Restauración a los ajustes predeterminados
La opción de restauración establece todos los parámetros del programa a los ajustes
predeterminados.
1. Pulse MENU (Menú).
2. Desplácese a PROGRAM SETUP>RESTORE (Configuración del programa>Restauración).
3. Seleccione YES (Sí).
Iniciar o parar un programa
Use la tecla RUN/HALT PROGRAM (Ejecutar/Detener programa) para iniciar o parar un programa.
El usuario puede definir hasta 12 horas/fechas de inicio/parada. Para poder modificar la
configuración de un sistema o programa hay que detener antes el programa.
1. Pulse RUN/HALT PROGRAM (Ejecutar/Detener programa).
2. Seleccione una de las opciones.
Opción Descripción
Start (Iniciar) Inicia el programa cargado actualmente. La indicación de estado pasa a
RUNNING (En ejecución).
Halt (Detener) Detiene temporalmente la ejecución del programa. La indicación de estado pasa
a HALTED (Detenido).
Resume (Reanudar) El programa reanuda su ejecución desde el punto en el que se detuvo. La
indicación de estado pasa a RUNNING (En ejecución).
Start over (Reiniciar) El programa comienza desde el principio. La indicación de estado pasa a
RUNNING (En ejecución).
End program (Terminar
el programa)
Detiene el programa que se está ejecutando actualmente. La indicación de
estado pasa a COMPLETE (Finalizado).
72 Español
Operación manual
Use la operación manual para recoger una muestra puntual, mover el brazo del distribuidor u operar
la bomba. Si desea obtener más información, consulte la versión ampliada de este manual.
Calibración del volumen
El volumen de la muestra se puede calibrar de forma automática con el sensor de líquido o de forma
manual. Cuando el sensor de líquido se usa con el fin de calibrar el volumen, queda ajustado para
medir con exactitud todos los volúmenes programados. Si el sensor de líquido está desactivado, el
volumen para las muestras, los lavados y las aguas de tormenta se deben calibrar manualmente.
En las aplicaciones típicas, se puede usar agua corriente para calibrar el volumen. Si la composición
de la muestra difiere significativamente de la composición de las muestras de agua típicas, calibre el
tomamuestras usando el líquido que se vaya a recoger. Se deben detener todos los programas
antes de efectuar la calibración.
Calibración del sensor de líquido
La precisión del sensor de líquido varía según el tipo de líquido que se esté muestreando. Por
ejemplo, el volumen de una muestra muy turbia puede no ser tan exacto como el del agua limpia. El
sensor de líquido se puede calibrar con el agua de muestreo para mejorar la precisión del volumen.
1. Coloque los tubos de admisión dentro del agua de muestreo.
2. Pulse MENU (Menú).
3. Para comprobar que el sensor de líquido está activado desplácese a PROGRAM
SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuración del
programa>Modificar selección>Sensor de líquido>Activar).
4. Vaya a SYSTEM SETUP>LIQ SENSOR CAL (Configuración del sistema>Calibración del sensor
de líquido).
5. Seleccione PERFORM CAL (Efectuar calibración).
6. Seleccione START (Iniciar). La bomba funcionará a la inversa para purgar el tubo de admisión. A
continuación, la bomba funcionará en dirección hacia delante.
7. Observe el caudal de muestra del tubo de admisión para comprobar que la muestra pasa por el
sensor de líquido.
8. Cuando la muestra pase a través del sensor de líquido y no haya burbujas, pulse la tecla STOP
(Detener). El tubo de admisión se purga automáticamente. La calibración del sensor de líquido
se ha efectuado.
9. Seleccione DONE (Listo) para salir o REPEAT (Repetir) para volver a realizar la calibración.
Restauración a los valores de calibración predeterminados
Es posible hacer que la calibración del volumen del sensor de líquido vuelva a tener los ajustes
predeterminados de fábrica.
1. Para comprobar que el sensor de líquido está activado desplácese a PROGRAM
SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuración del
programa>Modificar selección>Sensor de líquido>Habilitar).
2. Pulse VOLUME CALIBRATION (Calibración del volumen) y seleccione RESET CAL
(Restablecer calibración). La calibración del volumen está especificada según los ajustes de
fábrica. Aparecerá el mensaje “value has been reset to 0” (el valor se ha restablecido a cero).
Calibración del volumen de la muestra con el sensor de líquido
Calibre el volumen de la muestra con el sensor de líquido activado para ajustar levemente los
volúmenes de muestra.
Español
73
1. Para comprobar que el sensor de líquido está activado desplácese a PROGRAM
SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuración del
programa>Modificar selección>Sensor de líquido>Activar).
2. Introduzca los tubos de admisión en el agua de muestro o en agua corriente.
Nota: Para obtener una mejor precisión, use el agua de muestreo.
3. Quite los tubos del lado de salida de la bomba e introdúzcalos en una probeta.
4. Pulse VOLUME CALIBRATION (Calibración del volumen) y seleccione CALIBRATION
(Calibración).
5. Seleccione START (Iniciar). La bomba funcionará a la inversa para purgar el tubo de admisión. A
continuación, la bomba funcionará en dirección hacia delante y recogerá el volumen de la
muestra especificado en el programa de muestreo. La bomba funcionará a la inversa para purgar
el tubo de admisión.
6. Cuando se haya recogido la muestra, seleccione DONE (Listo). Compare el volumen de la
muestra de la probeta con el volumen de la muestra introducido en el menú de configuración del
programa.
7. Si el volumen que fue recogido difiere del volumen de muestra indicado en el programa actual,
especifique el volumen que fue recogido realmente. Pulse ENTER (Intro). El sensor se ajusta
para medir con precisión todos los volúmenes programados.
8. Seleccione START (Iniciar) para comenzar el programa de muestreo o pulse CANCEL
(Cancelar) para salir del menú de calibración del volumen.
Calibración del volumen de la muestra de forma manual
Si el sensor de líquido está desactivado, hay que calibrar todos los volúmenes de forma manual y
por tiempo. El volumen de muestra se calibra para el volumen especificado en el programa actual. Si
se cambia el volumen de la muestra en el programa, habrá que volver a calibrar el volumen de la
muestra para el nuevo valor.
1. Para comprobar que el sensor de líquido está desactivado desplácese a PROGRAM
SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuración del
programa>Modificar selección>Sensor de líquido>Desactivar).
2. Introduzca los tubos de admisión en el agua de muestro o en agua corriente.
3. Quite los tubos del lado de salida de la bomba e introdúzcalos en una probeta.
4. Pulse VOLUME CALIBRATION (Calibración del volumen). Si aparece más de un volumen,
seleccione uno de los volúmenes para realizar la calibración.
5. Seleccione START (Iniciar). La bomba funcionará a la inversa para purgar el tubo de admisión. A
continuación, la bomba funcionará en dirección hacia delante y comenzará a recoger la muestra.
6. Pare la bomba en el volumen o lugar de lavado seleccionado:
Volumen de la muestra: pulse STOP (Detener) cuando se haya recogido el volumen que se
especifica en el programa principal.
Lavado: pulse STOP (Detener) cuando el líquido de los tubos de admisión llegue al detector
de líquido.
Volumen de tormenta: pulse STOP (Detener) cuando se haya recogido el volumen que se
especifica en el programa de aguas de tormenta.
7. Seleccione DONE (Listo) para pasar a la siguiente calibración del volumen o pulse REPEAT
(Repetir) para volver a efectuar la calibración.
8. Cuando se hayan calibrado todos los volúmenes, pulse DONE (Listo).
9. Conecte los tubos de salida a la conexión del tubo del tomamuestras.
10. Seleccione START (Iniciar) para comenzar el programa de muestreo o CANCEL (Cancelar) para
salir.
74
Español
Verificación del volumen de la muestra
Para comprobar la precisión del volumen de la muestra, tome una muestra al azar. No vuelva a la
calibración para verificar el volumen, porque la compensación del volumen se restablece a 0 al inicio
de la calibración.
1. Pulse MANUAL OPERATION (Funcionamiento manual).
2. Seleccione GRAB SAMPLE (Muestra al azar).
3. Coloque los tubos de admisión dentro del agua de muestreo.
4. Quite los tubos del lado de salida de la bomba e introdúzcalos en una probeta.
5. Si el sensor de líquido está activado, introduzca el volumen que va a verificar.
6. Si el sensor de líquido está desactivado, introduzca el volumen que se especifica en el programa.
7. Pulse ENTER (Intro). Se inicia el ciclo de bomba (purga-muestra-purga).
8. Compare el volumen que se ha recogido en la probeta con el valor del volumen que se esperaba.
Si el volumen recogido no es el correcto, vuelva a efectuar la calibración del volumen.
Mantenimiento
P E L I G R O
Peligros diversos. Sólo el personal cualificado debe realizar las tareas descritas en esta sección del
documento.
P E L I G R O
Peligro de electrocución. Retire la alimentación eléctrica del instrumento antes de realizar actividades
de mantenimiento o reparación.
A D V E R T E N C I A
Exposición a riesgo biológico. Siga protocolos de manipuleo seguros mientras esté en contacto con
las botellas y los componentes del tomamuestras.
A D V E R T E N C I A
Peligros diversos. El técnico debe asegurarse de que el equipo funciona de forma segura y correcta
después de los procedimientos de mantenimiento.
Limpieza del instrumento
P R E C A U C I Ó N
Peligro de incendio. No utilice agentes inflamables para limpiar el instrumento.
A V I S O
No limpie el calefactor del compartimiento del controlador con ningún tipo de líquido.
Si el controlador y la bomba no se limpian bien solo con agua, desconecte el controlador y aléjelo
del muestreador. Espere el tiempo suficiente a que el controlador y la bomba se sequen antes de
volver a instalar las piezas y ponerlas en funcionamiento de nuevo.
Limpie el tomamuestras de la siguiente manera:
Carcasa del muestreador—Limpie la superficie interna y externa con un paño húmedo y
detergente suave. No utilice limpiadores abrasivos ni disolventes.
Español
75
Botellas para muestras—Limpie las botellas con un cepillo, agua y un detergente suave.
Enjuáguelas con agua corriente y, a continuación, con agua destilada. Si es necesario, utilice un
autoclave para limpiar las botellas de vidrio.
Cambio del desecante
Un cartucho desecante en el controlador absorbe la humedad y evita la corrosión. Controle el color
del desecante a través de la ventana (Figura 12). El desecante en buen estado es de color amarillo.
Cuando el color sea verde, cambie el desecante.
1. Desatornille y quite el cartucho de desecante (Figura 12).
2. Quite el tapón y tire el desecante gastado.
3. Llene dicho tubo con desecante nuevo.
4. Coloque el tapón.
5. Aplique grasa a la junta tórica.
6. Coloque el tubo de desecante en el controlador.
Figura 12 Cartucho desecante
1 Tapa 3 Junta tórica
2 Tubo con desecante 4 Ventana para observar el desecante
Mantenimiento de la bomba
P R E C A U C I Ó N
Riesgo de opresión. Retire la alimentación eléctrica del instrumento antes de realizar actividades de
mantenimiento o reparación.
Sustitución de los tubos de la bomba
A V I S O
Si se utilizan tipos distintos a los que suministra el fabricante puede provocarse un desgaste excesivo en las
piezas mecánicas o un mal rendimiento de la bomba.
Compruebe si los tubos de la bomba están desgastados en la zona donde los rodillos rozan con los
tubos. Sustituya los tubos cuando aparezca el mensaje de recordatorio o cuando estos muestren
indicios de desgaste.
Requisitos previos:
Tubos de la bomba: precortados o a granel, 4,6 m o 15,2 m (15 pies o 50 pies)
1. Desconecte la alimentación del controlador.
2. Si se utilizan tubos a granel, córtelos y agregue los puntos de alineación. Consulte el manual
completo en la página web del fabricante.
3. Quite los tubos de la bomba, como se muestra en los siguientes pasos ilustrados.
76
Español
4. Limpie los residuos de silicona del interior del alojamiento de la bomba y de los rodillos.
5. Coloque los nuevos tubos de la bomba, como se muestra en los siguientes pasos ilustrados.
Español 77
Limpieza del rotor
Limpie el rotor, los encauzadores del tubo de la bomba y el alojamiento de la bomba con un
detergente suave. Consulte el apartado Sustitución de los tubos de la bomba en la página 76 y los
siguientes pasos ilustrados.
Sustitución del tubo del brazo del distribuidor
El brazo del distribuidor se desplaza por encima de cada botella durante el muestreo con múltiples
botellas. Sustituya el tubo del brazo del distribuidor cuando el tubo esté desgastado. Asegúrese de
usar el tubo correcto para el distribuidor y el brazo del distribuidor correctos.
Nota: El tubo del distribuidor es diferente del tubo de la bomba. Si se instala el tubo de la bomba en el conjunto del
distribuidor, puede dañar el distribuidor. Además, se pueden perder muestras porque resulta difícil mover el brazo
del distribuidor.
1. Retire el tubo del brazo del distribuidor y del techo de la sección central.
2. Introduzca el nuevo tubo en el brazo del distribuidor. Asegúrese de que el tubo está alineado
conforme a las medidas que se muestran en la Figura 13.
3. Introduzca el otro extremo del tubo en la conexión del techo de la sección central.
4. Lleve a cabo una prueba de diagnóstico del distribuidor para comprobar que el funcionamiento
es correcto.
Figura 13 Conjunto distribuidor
1 Boquilla 3 Motor del distribuidor 5 Tubería del distribuidor
2 Brazo del distribuidor 4 Eje
78 Español
Sustitución de fusibles
El fusible no es una pieza que pueda reparar el usuario. Póngase en contacto con el servicio de
asistencia técnica. Un fusible fundido podría indicar que hay un problema en el muestreador y que
se necesita mantenimiento.
Español 79
Índice
Especificações na página 80 Operação na página 94
Informações gerais na página 82 Manutenção na página 100
Instalação na página 85
Informações adicionais
Você poderá encontrar informações adicionais no website do fabricante.
Especificações
As especificações estão sujeitas a alterações sem aviso prévio.
Amostrador Portátil SD900
Especificação Base padrão Base compacta Base de composto
Dimensões Diâmetro: 50,5 cm (19,9 pol.)
Altura: 69,4 cm (27,3 pol.)
Diâmetro: 44,1 cm
(17,4 pol.)
Altura: 61 cm (24 pol.)
Diâmetro: 50,28 cm
(19,8 pol.)
Altura: 79,5 cm
(31,4 pol.)
Peso 15 kg (35,6 lb) com garrafas de
polietileno (24x)
14,8 kg (32,6 lb) com recipiente de
polietileno de 10 L (2,5 gal) (1x)
12,2 kg (27 lb) com garrafas
de polietileno de 575 mL
(19,44 oz) (24x)
12,9 kg (28,3 lb) com
recipiente de polietileno de
10 L (2,5 gal) (1x)
15 kg (36 lb) com
garrafas de polietileno
de 950 mL (32,12 oz)
(12x)
Caixa ABS resistente a impacto, construção de 3 seções, base com paredes duplas com 2,54 cm
(1 pol.) isolamento — contato direto da garrafa com gelo.
Temperatura da
amostra
0 a 60 °C (32 a 140 °F)
Filtros Aço inoxidável 316 em tamanho padrão, alta velocidade ou perfil baixo para aplicativos de
profundidade rasa e Teflon
®
ou aço inoxidável 316 em tamanho padrão
Tubulação de
entrada de
amostra
9,5 mm (3/8 pol.) Polietileno de vinil I.D. ou revestido de Teflon
Capacidade da
garrafa de
amostra
Garrafas de polietileno de 1 L
(0,26 gal) e/ou de vidro de 350 mL
(11,83 oz) (24x)
Garrafas de polietileno de 2,3 L
(0,6 gal.) e/ou de vidro de 1,9 L
(0,5 gal.) (8x)
Garrafas de polietileno de 3,8 L
(1 gal.) e/ou de vidro de 3,8 L
(1 gal.) (4x)
Garrafas de polietileno de 3,8 L
(1 gal.) e/ou de vidro de 3,8 L
(1 gal.) (2x)
Recipiente composto de polietileno
de 21 L (5,5 gal) ou 15 L (4 gal) ou
polietileno de 20 L (5,25 gal) ou
polietileno de 10 L (2,5 gal) ou
vidro de 10 L (2,5 gal) (1x)
Garrafas de polietileno de
575 mL (19,44 oz) (24x)
Garrafas de vidro de
950 mL (32,12 oz) (8x)
Garrafa de polietileno de
10 L (2,5 gal) (1x)
Garrafa de vidro de 10 L
(2,5 gal) (1x)
Garrafa de polietileno
de 21 L (5,5 gal) (1x)
80 Português
controlador SD
Especificação Detalhes
Dimensões (L x A x P) 29,2 x 17,1 x 26.4 cm (11½ x 6¾ x 10
3
/
8
pol.)
Peso 4.2 kg (9.26 lb)
Caixa Mescla PC/ABS, NEMA 4X, 6, IP 67
Alimentação elétrica 15 Vcd alimentado por uma fonte de alimentação 8754500; 15 Vcd
alimentado por uma fonte de alimentação integral
Proteção de sobrecarga Fusível de linha CD de 7 A para a bomba
Temperatura de operação 0 a 50°C (32 a 122°F)
Temperatura de armazenamento –30 a 60°C (-22 a 140°F)
Umidade de armazenamento/operação Condensação de 100%
Bomba Alta velocidade peristáltica, com cilindros de Nylatron montados em
molas
Caixa da bomba IP37
Tubo da bomba Silicone com DI de 9,5 mm x DE de 15,9 OD mm (
3
/
8
pol. de DI x
5
/
8
pol. de DE)
Vida da tubulação da bomba 20.000 ciclos de amostra com: volume de amostra de 1 l (0,3 gal.),
1 lavagem, intervalo com ritmo de 6 minutos, 4,9 m (16 pés) de tubo
de entrada de
3
/
8
pol., 4,6 m (15 pés) de elevação vertical,
temperatura de amostra de 21 °C (70 °F)
Elevação vertical de amostra Máximo 8,5 m (28 pés) para: 8,8 m (29 pés) de tubo de entrada de
vinil de
3
/
8
pol. ao nível do mar a 20–25 °C (68–77 °F)
Taxa de fluxo da bomba 4,8 l/min (1,25 gpm) na elevação vertical de 1 m (3 pés) com tubo de
entrada de
3
/
8
pol. típico
Volume de amostra Programável em incrementos de 10 ml (0,34 oz) de 10 a 10.000 ml
(3,38 oz. a 2,6 gal.)
Repetibilidade de volume de amostra
(típica)
±5% de 200 ml volume da amostra usando detecção de líquido sem
calibragem com: com suspensão vertical de 4,6 m (15 pés), 4,9 m
(16 pés) de tubo de entrada de vinil de
3
/
8
pol., garrafa única, válvula
de garrafa cheia em temperatura ambiente e elevação de 1524 m
(5000 pés)
Precisão de volume de amostra (típica) ±10% de volume da amostra de 200 ml usando detecção de líquido
sem calibragem com: com suspensão vertical de 4,6 m (15 pés), 4,9 m
(16 pés) de tubo de entrada de vinil de
3
/
8
pol. garrafa única, válvula
de garrafa cheia em temperatura ambiente e elevação de 1524 m
(5000 pés)
Modos de amostragem Passo: fixo-tempo, fixo-fluxo, variável-tempo, variável-fluxo, volume
variável de tempo constante (CTVV).
Consulte a amostragem Volume variável de tempo constante (CTVV)
na versão estendida deste manual.
Distribuição: composto de frasco único, composto de vários frascos,
separado de vários frascos, frascos por amostra, amostras por frasco,
combinação de frascos por amostra, amostras por frasco
Modos de execução Contínuo ou não-contínuo com número de amostras inserido pelo
usuário
Programas múltiplos Armazena até três programas de amostragem
Velocidade de transferência (típica) 0,9 m/s (2,9 pés/s) com: elevação vertical de 4,6 m (15 pés), 4,9 m
(16 pés) de tubulação de entrada de vinil de
3
/
8
pol. 21 °C (70 °F) e
elevação de 1524 m (5000 pés)
Português 81
Especificação Detalhes
sensor de líquido Ultrasônico. Corpo: Ultem
®
NSF ANSI aprovado pelo padrão 51, de
acordo com a USP Classe VI
histórico de amostra Até 510 registros
Purga de ar Ar purgado automaticamente antes e depois de cada amostra. A
duração compensa automaticamente as variações no comprimento
dos tubos de entrada.
Registro de eventos 510 registros
Conexões
Materiais molhados Aço inoxidável, polietileno, Teflon, Ultem, silicone
Comunicações RS232, Modbus, SDI-12
Garantia 1 ano
Informações gerais
Em hipótese alguma o fabricante será responsável por danos diretos, indiretos, especiais,
incidentais ou consequenciais resultantes de qualquer defeito ou omissão neste manual. O
fabricante reserva-se o direito de fazer alterações neste manual e nos produtos aqui descritos a
qualquer momento, sem aviso ou obrigação. As edições revisadas podem ser encontradas no site
do fabricante.
Informações de segurança
A V I S O
O fabricante não é responsável por quaisquer danos devido ao uso ou aplicação incorreta deste produto,
incluindo, sem limitação, danos diretos, acidentais ou consequenciais, e se isenta desses danos à extensão total
permitida pela lei aplicável. O usuário é unicamente responsável por identificar riscos críticos de aplicação e por
instalar os mecanismos apropriados para proteger os processos durante um possível mau funcionamento do
equipamento.
Leia todo o manual antes de tirar da embalagem, montar ou operar esse equipamento. Preste
atenção a todas as declarações de perigo e cuidado. Caso contrário, o operador poderá sofrer
ferimentos graves ou o equipamento poderá ser danificado.
Certifique-se de que a proteção oferecida por este equipamento não seja afetada. Não use nem
instale este equipamento de nenhuma outra forma além da especificada neste manual.
Uso de informações de risco
P E R I G O
Indica uma situação potencial ou iminentemente perigosa que, se não for evitada, resultará em morte ou lesão
grave.
A D V E R T Ê N C I A
Indica uma situação potencialmente perigosa que, se não for evitada, pode resultar em morte ou ferimento grave.
C U I D A D O
Indica uma situação potencialmente perigosa que pode resultar em ferimento leve a moderado.
A V I S O
Indica uma situação que, se não evitada, pode causar danos ao instrumento. Informações que necessitam de
uma ênfase especial.
82 Português
Avisos de precaução
Leia todas as etiquetas e rótulos fixados no instrumento. Caso não sejam observadas, podem
ocorrer lesões pessoais ou danos ao instrumento. Um símbolo no instrumento tem sua referência no
manual com uma medida preventiva.
Este é o símbolo de alerta de segurança. Acate todas as mensagens de segurança que seguem este
símbolo a fim de evitar lesões potenciais. Se o símbolo estiver no instrumento, consulte o manual de
instruções para obter informações sobre a operação ou segurança.
Este símbolo indica possível risco de pinçamento.
O equipamento elétrico marcado com este símbolo não pode ser descartado em sistemas de
descarte público ou doméstico europeus. Devolva equipamentos antigos ou no final da vida útil para
o fabricante para descarte, sem custo adicional para o usuário.
Certificação
Canadian Radio Interference-Causing Equipment Regulation (Regulamentação para
equipamentos de rádio causadores de interferência do Canadá), IECS-003, Classe A:
Os registros de testes de comprovação encontram-se com o fabricante.
Este aparelho digital Classe A atende a todos os requisitos de regulamentações canadenses sobre
equipamentos que causam interferências.
Cet appareil numèrique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC parte 15, limites Classe "A"
Os registros de testes de comprovação encontram-se com o fabricante. O dispositivo está em
conformidade com a Parte 15 das Regras da FCC. A operação está sujeita às seguintes condições:
1. O equipamento não deve causar interferência prejudicial.
2. O equipamento deve aceitar todas as interferências recebidas, inclusive interferências que
podem causar funcionamento indesejado.
Alterações ou modificações a este equipamento não aprovadas expressamente pela parte
responsável pela conformidade podem anular a autoridade do usuário de operar o equipamento.
Este equipamento foi testado e está em conformidade com os limites de dispositivo digital Classe A,
de acordo com a Parte 15 das Regras da FCC. Esses limites foram estabelecidos para proporcionar
uma razoável proteção contra interferências nocivas quando o equipamento for operado em
ambientes comerciais. Este equipamento gera, utiliza e pode irradiar energia de rádiofrequência e,
se não instalado e usado de acordo com o manual de instruções, pode causar interferências
prejudiciais às comunicações de rádio. É provável que o funcionamento deste equipamento em área
residencial possa causar interferência indesejada, caso em que o usuário será solicitado a corrigir a
interferência por conta própria. As seguintes técnicas podem ser usadas para reduzir problemas de
interferência:
1. Desconecte o equipamento de sua fonte de alimentação para verificar se ele é ou não a origem
da interferência.
2. Se o equipamento está conectado à mesma tomada do dispositivo que está sofrendo
interferência, conecte o equipamento a uma tomada diferente.
3. Afaste o equipamento do dispositivo que estiver recebendo a interferência.
4. Reposicione a antena de recebimento do dispositivo que está sofrendo interferência.
5. Tente algumas combinações das opções acima.
Português
83
Visão geral do produto
P E R I G O
Riscos químicos ou biológicos. Se esse instrumento for usado para monitorar um processo de
tratamento e/ou sistema de alimentação química para o qual existam limites de regulamentação e
requisitos de monitoramento relacionados à saúde pública, à produção ou ao processamento de
alimentos ou bebidas, é responsabilidade do usuário deste instrumento conhecer e cumprir as
regulamentações aplicáveis e ter mecanismos suficientes e apropriados para obter conformidade com
as regulamentações aplicáveis no caso de mau funcionamento do instrumento.
C U I D A D O
Perigo de incêndio. Este produto não foi projetado para uso com líquidos inflamáveis.
O Amostrador Portátil SD900 coleta amostras de líquidos em intervalos específicos e mantém as
amostras em garrafas ou recipientes. Use o amostrador para uma variedade de aplicações aquosas
e para poluentes tóxicos e sólidos suspensos. Configure o amostrador com diferentes retentores,
garrafas e recipientes. Consulte Figura 1.
Figura 1 Visão geral do produto
1 Base compacta 6 Conexões do controlador
2 Base isolada padrão 7 Bomba
3 Seção central 8 Controlador
4 Fonte de energia 9 sensor de líquido
5 Cobertura superior 10 Recipiente de base isolada padrão para 21 L
(5,5 gal)
Componentes do produto
Certifique-se de que todos os componentes foram recebidos. Consulte Figura 2. Se houver itens
ausentes ou danificados, entre em contato imediatamente com o fabricante ou com um
representante de vendas.
84
Português
Figura 2 Componentes do produto
1 Base (padrão, compacta ou composta) 6 Carregador de bateria (opcional)
2 Componentes para uma opção de garrafa única
(garrafa e suporte podem mudar)
7 Suprimento de força CA (opcional)
3 Seção central com controlador 8 Filtro
4 Bateria (opcional) 9 Tubulação de entrada, coberta de vinil ou teflon
5 Cobertura superior 10 Componentes para uma opção de múltiplas
garrafas (as garrafas e os retentores podem
mudar)
Instalação
P E R I G O
Vários perigos. Somente pessoal qualificado deve realizar as tarefas descritas nesta seção do manual.
Precauções em espaços confinados
P E R I G O
Perigo de explosão. Treinamento em testes pré-entrada, ventilação, procedimentos de entrada,
procedimentos de evacuação/resgate e práticas de trabalho de segurança são necessárias antes de
entrar em espaços confinados.
As informações a seguir são fornecidas para ajudar os usuários a entenderem os perigos e os riscos
associados com a entrada em espaços confinados.
Em 15 de abril de 1993, a decisão final da OSHA sobre o CFR 1910.146, Autorização Requerida
para Espaços Confinados, se tornou lei. Este padrão afeta diretamente mais de 250.000 locais
industriais nos EUA e foi criado para proteger a saúde e a segurança dos trabalhadores em espaços
confinados.
Definição de um espaço confinado:
Um espaço confinado é qualquer local ou recinto que apresente (ou tenha potencial imediato para
apresentar) uma ou mais das seguintes condições:
Uma atmosfera com uma concentração de oxigênio menor que 19,5% ou maior que 23,5% e/ou
uma concentração de sulfeto de hidrogênio (H
2
S) que seja maior que 10 ppm.
Uma atmosfera que possa ser inflamável ou explosiva devido a gases, vapores, névoas, poeira ou
fibras.
Português
85
Materiais tóxicos que, mediante contato ou inalação, podem causar lesões, danos à saúde ou
morte.
Os espaços confinados não são feitos para ocupação humana. Os espaços confinados têm uma
entrada restrita e contêm riscos conhecidos ou potenciais. Exemplos de espaços confinados incluem
câmaras subterrâneas, chaminés, tanques, subterrâneos de troca e outros locais semelhantes.
Os procedimentos de segurança padrão devem sempre ser obedecidos antes da entrada nos
espaços confinados e/ou locais onde possam estar presentes gases perigosos, vapores, névoas,
poeiras ou fibras. Antes de entrar em um local confinado, encontre e leia todos os procedimentos
relacionados à entrada em um espaço confinado.
Instalação mecânica
Diretrizes da instalação no local
P E R I G O
Risco de explosão. O instrumento não está aprovado para ser instalado em localizações perigosas.
Consulte as diretrizes que se seguem para a avaliação da localização do site.
Obedeça a todas as precauções de segurança se o amostrador for instalado em um local
confinado. Consulte Precauções em espaços confinados na página 85.
Certifique-se de que a temperatura no local está no intervalo específico. Consulte Especificações
na página 80.
Instale o amostrador em uma superfície nivelada ou pendure o amostrador com o suporte de
suspensão, a alça de suporte ou a barra de abertura. Consulte Instale o amostrador em um bueiro
na página 86 e a documentação da instalação aplicável.
Esteja o mais próximo possível da origem da amostra para reduzir a demora da análise. Consulte
Instale a tubulação do amostrador na página 89.
Para obter as limitações sobre a velocidade do transporte e máxima elevação vertical, consulte
Especificações na página 80.
Instale o amostrador em um bueiro
Instale o amostrador acima da água de amostra em um bueiro. Instale o amostrador com uma barra
de abertura ou uma alça de suporte. Instale a barra de abertura dentro do bueiro. A barra de
abertura é suportada pela pressão contra as paredes. A alça de suporte tem a mesma largura que a
tampa do bueiro. Instale a alça de suporte diretamente abaixo da tampa para obter suporte.Consulte
o manual expandido no site do fabricante.Consulte a documentação fornecida com os acessórios
para instalar o amostrador.
Instale o distribuidor ou o fechamento total da garrafa (opcional)
O conjunto do distribuidor ou o fechamento total da garrafa é normalmente instalado na fábrica.
Consulte a documentação do distribuidor ou do fechamento total da garrafa para a instalação.
Prepare o amostrador
Limpe as garrafas de amostra
Limpe as garrafas de amostra e as tampas com escova, água e detergente suave. Lave os
recipientes com água corrente e, em seguida, enxágue com água destilada.
Instale uma única garrafa
Use uma única garrafa para coletar uma amostra composta. Quando a garrafa estiver cheia, a
válvula de garrafa cheia para o programa de amostra.
1. Limpe as garrafas de amostra. Consulte Limpe as garrafas de amostra na página 86.
2. Instale a garrafa de amostra como mostrado na Figura 3.
86
Português
Figura 3 Instalação de garrafa única
1 Garrafa de polietileno, 10 L (2,5 gal) (1918) 6 Garrafa de polietileno, 21 L (5,5 gal) (6494)
2 Garrafa de vidro, 10 L (2,5 gal) (6559) 7 Base compacta (8975)
3 Suporte (1502) 8 Base isolada padrão (8976)
4 Garrafa de polietileno, 15 L (4 gal) (1367) 9 Garrafa de base isolada padrão para 21 L (5,5 gal)
(8561)
5 Garrafa de polietileno, 19 L (5 gal) (6498)
Instale a primeira garrafa para várias coleções de amostras
Use garrafas múltiplas para coletar amostras em garrafas separadas ou em mais do que uma
garrafa. O distribuidor move o tubo de amostra acima de cada garrafa. Instale as garrafas na base
do amostrador como indicado na Figura 4. Instale a primeira garrafa de amostra (número 1) abaixo
da etiqueta na base do amostrador. Instale as demais garrafas em números crescentes na direção
indicada pela etiqueta. Consulte Figura 5 na página 89 para obter um diagrama dos componentes
necessários.
Português
87
Figura 4 Instalação da garrafa número 1
1 Localização da garrafa número 1 (base compacta) 6 Faixas elásticas
2 Localização da garrafa número 1 para 24 garrafas 7 Garrafa de base isolada padrão para 21 L (5,5 gal)
3 Localização da garrafa número 1 para 8 garrafas 8 Base isolada padrão
4 Localização da garrafa número 1 para 2 ou
4 garrafas
9 Base compacta
5 Retentor
Instale várias garrafas
Quando várias garrafas estão instaladas, um braço distribuidor move o tubo de amostra sobre cada
garrafa. A coleta de amostra é automaticamente parada quando o número especificado de amostras
tiver sido coletado.
1. Limpe as garrafas de amostra. Consulte Limpe as garrafas de amostra na página 86.
2. Instale a primeira garrafa de amostra (número 1) abaixo da etiqueta na base do amostrador.
Consulte Instale a primeira garrafa para várias coleções de amostras na página 87.
3. Ordene as garrafas de amostra como mostrado na Figura 5. Para oito garrafas ou mais,
certifique-se de que colocar a primeira garrafa próxima ao indicador da garrafa um no sentido
horário.
4. Coloque o conjunto de garrafas no amostrador. Para oito garrafas ou mais, alinhe a grade nas
ranhuras da bandeja de garrafas.
88
Português
Figura 5 Instalação de garrafas múltiplas
1 Posicionador/retentor (2347) 8 Garrafas de vidro, 3,8 L (1 gal), (4x) (2216) ou
garrafas de polietileno, 3,8 L (1 gal), (4x) (2217)
2 Retentor (1422) 9 Garrafas de vidro, 1,9 L (0,5 gal), (8x) (1118) ou
garrafas de polietileno, 2,3 L (0,61 gal), (8x) (657)
3 Retentor (2190) 10 Garrafas de polietileno, 1 L (0,26 gal), (24x) (737)
4 Retentor (2189) 11 Garrafas de vidro, 350 mL (11,83 oz) (24x) (732)
5 Garrafas de vidro, 950 mL (32,12 oz) (8x) (2348) 12 Base compacta (8975)
6 Garrafas de polietileno, 575 mL (19,44 oz), (24x)
(1369)
13 Base padrão (8976)
7 Garrafas de vidro, 3,8 L (1 gal), (2x) (2214) ou
garrafas de polietileno, 3,8 L (1 gal), (2x) (2215)
Instale a tubulação do amostrador
Instale o filtro no meio do fluxo da amostra (não próximo à superfície ou ao fundo) para se certificar
de que uma amostra representativa seja coletada. Consulte Figura 6. Consulte Figura 7 para a
instalação do tubo de entrada.
1. Conecte a tubulação ao amostrador como mostrado na Figura 7.
Observação: Use o kit 2186 de conexão se a tubulação revestida de Teflon for usada.
2. Instale o tubo de entrada e o filtro no fluxo principal da fonte de amostra onde a água está
turbulenta e bem misturada.
Deixe o tubo de entrada o mais curto possível.
Mantenha o tubo de entrada em uma inclinação mais vertical possível para que seque
completamente entre uma amostra e outra.
Observação: Se não for possível manter uma inclinação vertical ou se o tubo for pressurizado, desative o
sensor de líquido. Calibre o volume de amostra manualmente. Consulte Calibre o volume de amostra
manualmente na página 100.
Certifique-se de que o tubo de entrada não esteja comprimido.
Português
89
Figura 6 Instalação de instrumento
1 Filtro 3 Tubo de entrada
2 Elevação vertical
Figura 7 Instalação do tubo de entrada
Instalação elétrica
Conexões do controlador
A D V E R T Ê N C I A
Risco de choque elétrico. Equipamento conectado externamente deve ter uma avaliação apropriada do
padrão de segurança do país.
A Figura 8 mostra as conexões que podem ser feitas ao controlador.
90
Português
Figura 8 Conectores do controlador
1 Dispositivo auxiliar 4 Opção de dispositivo SDI-12
2 Fonte de alimentação 5 Distribuidor/válvula de garrafa cheia
3 Comunicações seriais
Conecte um medidor de fluxo
Conecte um medidor de fluxo ao controlador para iniciar ou parar o amostrador quando o fluxo de
amostra ficar acima ou abaixo de um valor especificado.
Itens para coletar:
Cabo pleno multi-fim para medidores de fluxo Sigma (ou meio cabo 980 para o modelo de
medidor de fluxo 980)
Divisor opcional para conexões extras. É possível conectar dois ou mais divisores em série.
1. Conecte uma extremidade do cabo ao medidor de fluxo. Para o medidor de fluxo modelo 980,
consulte o manual do usuário do medidor de fluxo modelo 980.
2. Conecte a outra extremidade do cabo ao conector auxiliar do dispositivo no controlador.
Observação: Se o medidor de fluxo possuir cabo de 6 pinos, use o cabo adaptador de 6 a 7 pinos.
Conecte um PC ou rede de comunicações
Conecte um PC ou uma rede de comunicações ao controlador para transferir dados ou para
configurar o amostrador.
Itens para coletar:
Cabo serial, RS232 de 7 pinos para DB-9 (consulte a Figura 9). Conexões: B no 5 (sinal terra); D
no 3 (RCD); F no 2 (TXD); G (terra).
1. Conecte uma extremidade do cabo serial ao dispositivo ou rede de comunicação.
2. Conecte a outra extremidade do cabo ao conector serial de comunicações no controlador.
Figura 9 Cabo serial de comunicações
Português 91
Conecte o amostrador à alimentação
P E R I G O
Risco de choque elétrico. Se este equipamento for usado ao ar livre ou em locais potencialmente
úmidos, um dispositivo contra Falhas de Aterramento (GFCI/GFI, Ground Fault Circuit Interrupt) deve
ser usado para conectar o equipamento à sua fonte principal de energia.
P E R I G O
Perigo de incêndio. Instale um disjuntor de 15 A na linha de alimentação. Um disjuntor pode ser o
interruptor local, caso esteja bem próximo ao equipamento.
P E R I G O
Risco de eletrocussão. A conexão com aterramento protetor (PE) é obrigatória
A D V E R T Ê N C I A
Risco de eletrocussão. Certifique-se de que haja um fácil acesso para desconectar a energia local.
A D V E R T Ê N C I A
Risco de choque elétrico. A fonte de alimentação pode se sobreaquecer se o tempo entre os ciclos de
bombeamento for muito curto.
A V I S O
Certifique-se de que o período de ciclo da bomba não sobreaquecerá a fonte de alimentação. Consulte Calcule o
tempo de ciclo da bomba na página 93.
Conecte o amostrador à bateria ou a uma fonte de alimentação AC. Consulte as etapas ilustradas
seguintes.
92
Português
Calcule o tempo de ciclo da bomba
A bomba deve permanecer desligada por um período mínimo de tempo entre os ciclos de amostra,
para que a fonte de alimentação não sobreaqueça.
1. Calcule o tempo total de funcionamento contínuo da bomba durante um ciclo de amostra. Inclua
todos os estágios: pré-purgação, lavagem de entrada, amostra, tentativas de amostra e pós-
purgação.
2. Encontre o menor período de tempo em que a bomba deve ficar desligada entre ciclos de
amostra na Figura 10.
3. Certifique-se de que o ritmo ou intervalo de tempo do programa amostrador permita que a
bomba fique desligada entre ciclos de amostra pela quantidade mínima de tempo.
Figura 10 Períodos em que bomba fica ligada e desligada a 50 °C (122 ºF)
1 Tempo (em minutos) em que a bomba deve
permanecer desligada
3 Exemplo: se a bomba permanecer ligada
continuamente por 5 minutos, ela deve ficar
desligada por 15 minutos.
2 Tempo (em minutos) em que a bomba deve
permanecer ligada continuamente
Ligue a alimentação
Pressione o botão de ligação para ligar ou desligar. Certifique-se de que a fonte de alimentação
(alimentação AC ou alimentação da bateria) esteja corretamente instalada.
Interface do usuário e navegação
O teclado do controlador é mostrado na Figura 11. Use a tecla MENU para configurar os programas
de amostragem, para definir as configurações do controlador ou para realizar testes de diagnóstico.
Use as teclas de seta, ENTER e BACK (Voltar) para rolar pelo menu, fazer seleções e inserir
valores. Procure setas na parte inferior ou lateral da tela para saber quando mais opções estão
disponíveis.
Português
93
Figura 11 Teclado do controlador
1 POWER (Força) 5 MENU 9 STATUS
2 VOLUME CALIBRATION
(Calibração de volume)
6 ENTER 10 RUN/HALT PROGRAM
(Executar/interromper
programa)
3 STOP (Parar) 7 VOLTAR 11 MANUAL OPERATION
(Operação manual)
4 LED 8 TECLAS DE SETAS
Operação
Defina as configurações do sistema
1. Pressione MENU.
2. Vá até SYSTEM SETUP (Configuração do sistema).
3. Altere as opções aplicáveis.
Opção Descrição
Configuração de
horário/data
Ajusta o horário (formato 24 horas) e a data.
comunicação Seleciona a taxa de transmissão (19200, 38400, 57600 ou 115200) e o
protocolo (Modbus RTU ou ASCII) para a porta serial.
Configurar base Seleciona a base do amostrador (Refrigerado para todos os climas,
refrigerado, padrão portátil ou compacto portátil).
Idioma Seleciona o idioma do controlador.
Liq sensor cal (cal sensor
de liq)
Calibra o sensor líquido ou define como padrão.
Ajuste de contraste Ajusta o contraste da tela de LCD.
Ciclo de operação da
bomba
Altera o ciclo de operação da bomba. Faixa: de 50% a 100% (padrão 100%).
94 Português
Opção Descrição
Ajuste de senha Define uma senha para controlar o acesso aos menus de configuração do
programa e do sistema. Selecione ENABLE>YES (Habilitar>Sim) para
configurar uma nova senha ou ENABLE>NO (Habilitar>Não) para ativar a
senha atual. Se a senha estiver sendo definida pela primeira vez, use
900900 como senha atual.
Vida útil da tubulação Define um lembrete de quando a tubulação da bomba deve ser trocada. Vá
até ENABLE>RESET CYCLES (Habilitar>Redefinir ciclos) para redefinir a
contagem do ciclo da bomba como 0. Acesse ENABLE>CYCLE LIMIT
(Habilitar>Limite do ciclo) para alterar o número de ciclos da bomba a ser
usado para o lembrete. Vá até ENABLE>STATUS (Habilitar>Status) para
mostrar o número atual de contagens de ciclo da bomba.
SDI-12 (opcional) Configura a operação do dispositivo SDI–12. Essa opção é exibida somente
quando um dispositivo é encontrado.
Configure um programa de amostragem
Um programa de amostragem inclui todos os parâmetros necessários para coletar amostras
automaticamente. Colete as amostras em intervalos regulares ou quando o programa de
amostragem estiver completo.
1. Pressione MENU.
2. Acesse PROGRAM SETUP>MODIFY ALL(Configuração do programa>Modificar tudo). A
primeira tela de configuração será exibida.
3. Selecione ou insira valores para cada parâmetro. Use as teclas de seta para a esquerda e para a
direita para mover para a esquerda ou direita. Use as teclas de seta para cima e para baixo para
mudar um valor. Mude as unidades se necessário (gal. Ou mL) com as teclas de flechas.
Opção Descrição
garrafas Quantidade de garrafa: o número de garrafas do amostrador (1, 2, 4, 8 ou 24).
Válvula de GRF cheia: ativada/desativada. Quando ativada, o controlador verifica a
garrafa quanto às condições de transbordamento.
Volume da garrafa: a capacidade de volume de cada garrafa. Faixa: 50–65000 mL
(0,01 - 17,17 gal)
tubulação de
entrada
Comprimento da tubulação: o comprimento da tubulação de entrada do filtro até o
sensor de líquido. Faixa: 100–3000 cm ou 3–99 pés. Um comprimento preciso é
necessário para obter um volume de amostra preciso.
Tipo de tubo de entrada: o diâmetro e o material da tubulação de entrada (¼ pol. vinil,
3/8 pol. vinil ou 3/8 pol. Teflon).
Atraso do
programa
Ativado/desativado: quando ativado, o programa de amostragem inicia em um horário
ou número de contagem especificado. Data e hora: a data e a hora em que o programa
será iniciado (formato 24 horas). Contagens: o número de contagens ou pulsos de um
medidor de fluxo no qual o programa será iniciado (1 a 9999 contagens). Se o ritmo de
amostra for depois selecionado para horário, o atraso do programa será desabilitado.
Ritmo de
amostra/coleta
Baseado tempo—Consulte Coleção baseada em tempo na página 96.
Observação: Quando CTVV é usado:
O sensor de líquido está sempre habilitado.
A distribuição de amostras é composta.
O modo de execução não é contínuo.
A primeira amostra é obtida após o intervalo.
A fonte de fluxo é sempre uma entrada de 4 a 20 mA.
Baseado fluxo—Consulte Coleção baseada em fluxo na página 97.
Português 95
Opção Descrição
Distribuição de
amostra
Entregar amostras a todos>YES (Sim)
Cada amostra é entregue a todas as garrafas. Finalizar após última: o programa para
após a última amostra (1–999). Contínuo: o programa continua até ser parado
manualmente.
Entregar amostras a todos>NO (Não)
As amostras são entregues a um subconjunto de garrafas . Amostras/garrafa: o
número de amostras a ser coletado em cada garrafa (1–999). Garrafas/amostra: o
número de garrafas que conterão a mesma amostra.
sensor de líquido Habilitar ou desabilitar. Se desabilitado, o volume de amostra deverá ser calibrado pelo
tempo.
Volume de
amostra
O volume de cada amostra em ml (1 gal. = 3785,4 ml). Se for usado o modo de
garrafas/amostras, cada garrafa recebe um volume completo de amostras
(10–10.000 ml). Certifique-se de que o volume de amostra não exceda o volume de
garrafas. Os volumes de amostra são arredondados para os 10 mL mais próximos.
Lavagens de
entrada
O número de lavagens do tubo de entrada antes de uma amostra ser coletada (0–3).
Tentativas de
amostra
O número de tentativas de amostragem após a falha de cada amostragem (0–3).
Identidade do
local
O nome do local de amostragem (até 12 caracteres). A identidade do lugar será usada
como um nome pré-definido se o programa for armazenado.
amostragem
avançada
Vai até o menu de amostragem avançada. Se necessário, configure as opções no
menu Advanced Sampling (Amostragem Avançada) menu.
Executar
programa
A criação do programa de amostragem básica é concluída. O amostrador solicita ao
usuário para iniciar ou cancelar o programa.
Coleção baseada em tempo
1. Selecione ou insira os valores para cada opção.
Opção Descrição
Intervalo de ritmo Coleta amostras em intervalos de tempo regulares, em horas e minutos
(0:01–999:00).
Tome a primeira
amostra
O programa é iniciado imediatamente ou após o primeiro intervalo de tempo.
Volume variável Especifique se os volumes de amostra variarão (baseados em fluxo). Em caso
afirmativo, consulte Use volume variável na página 96. Em caso negativo,
consulte Não use volume variável na página 97.
Use volume variável
A recolha de amostra é volume variável de tempo constante (CTVV).
1. Selecione ou insira os valores para cada opção.
Opção Descrição
Selecionar unidade Selecione as unidades de fluxo a serem usadas.
Entrada Map 4 mA Insira a taxa de fluxo em concordância com a entrada de 4 mA.
Entrada Map 20 mA Insira a taxa de fluxo em concordância com a entrada de 20 mA.
Taxa de fluxo média Especifique a taxa de fluxo média para o site (1–999.999). As unidades são
especificadas pelas unidades selecionadas acima.
96 Português
Opção Descrição
Volume de tempo desejado Especifique o volume total a ser coletado durante o programa de amostra
(10–10.000 ml).
Período de coleta Especifique o período de coleta total em horas e minutos (0:01–999:00).
Volume mínimo da amostra Se o volume de fluxo não for suficiente para coletar este volume mínimo, a
amostra será ignorada (10–10.000 ml).
Não use volume variável
1. Selecione ou insira os valores para a opção.
Opção Descrição
Tome a primeira amostra Selecione se o programa iniciará imediatamente ou após ter passado o primeiro
intervalo de tempo.
Coleção baseada em fluxo
Especifique a fonte de fluxo para amostragem baseada em fluxo:
Contagens—Consulte Use contagens para a fonte de fluxo na página 97.
Entrada de 4 a 20 mA—Consulte Use entrada de 4 a 20 mA para a fonte de fluxo na página 97.
Use contagens para a fonte de fluxo
1. Selecione ou insira os valores para cada opção.
Opção Descrição
Tomar amostra cada Colete amostras a intervalos de fluxo regulares, em contagens
(1–9.999 contagens).
Tempo limite Colete uma amostra se o volume de fluxo for muito baixo.
Hora O tempo máximo entre as amostras (0:01–999:00). O temporizador começa
novamente após a coleta de cada amostra.
Tome a primeira
amostra
O programa é iniciado imediatamente ou após o primeiro intervalo de tempo.
Use entrada de 4 a 20 mA para a fonte de fluxo
1. Selecione ou insira os valores para cada opção.
Opção Descrição
Selecionar unidade Seleciona as unidades de fluxo a serem usadas.
Entrada Map 4 mA Insira a taxa de fluxo em concordância com a entrada de 4 mA.
Entrada Map 20 mA Insira a taxa de fluxo em concordância com a entrada de 20 mA.
Tomar amostra cada Insira o fluxo total após o qual a amostra deve ser extraída.
Cancelar tempo Selecione habilitar para forçar que uma amostra seja coletada se o volume de fluxo
estiver excepcionalmente baixo.
Português 97
Opção Descrição
Hora O tempo máximo entre as amostras (0:01–999:00). O temporizador começa
novamente após a coleta de cada amostra.
Volume de amostra Insira o volume a ser coletado por amostra. Se o modo de garrafas/amostras for
selecionado, cada garrafa recebe um volume completo de amostras
(100–10.000 mL). Esta opção é ignorada quando o volume variável é selecionado.
Como restaurar as configurações padrão
A opção de restauração define todos os parâmetros de programa com as configurações padrão.
1. Pressione MENU.
2. Vá até PROGRAM SETUP>RESTORE (Configuração do programa>Restaurar).
3. Selecione YES (Sim).
Iniciar ou parar um programa
Use a tecla RUN/HALT PROGRAM (Executar/interromper programa) para iniciar ou parar um
programa. Podem ser definidos pelo usuário até 12 vezes/datas de iniciar/parar. Um programa deve
ser parado antes de ser possível modificar o programa ou a configuração do sistema.
1. Pressione RUN/HALT PROGRAM (Executar/interromper programa)
2. Selecionar uma das opções de saída.
Opção Descrição
Botão de inicialização Inicia o programa carregado atualmente. O estado muda para RUNNING (Em
execução).
Interromper Para o programa temporariamente. O estado muda para HALTED (Interrompido).
Reiniciar O programa reinicia a partir do ponto no qual foi interrompido. O estado muda para
RUNNING (Em execução).
Começar novamente O programa inicia do princípio. O estado muda para RUNNING (Em execução).
Finalizar programa Interrompe o programa atualmente em execução. O estado do programa muda
para COMPLETE (Concluído).
operação manual
Use a operação manual para coletar uma amostra de tanque, mover o braço distribuidor, ou operar
a bomba. Para obter mais informações, consulte a versão completa deste manual.
Calibração de volume
O volume de amostra pode ser calibrado automaticamente com o sensor de líquido ou
manualmente. Quando um sensor de líquido for usado para calibragem de volume, o sensor será
ajustado para medir com precisão todos os volumes programados. Se o sensor de líquido estiver
desabilitado, o volume para amostras, a lavagens e água pluvial devem ser calibrados
manualmente.
Para aplicações típicas, pode ser usada água de torneira para a calibração de volume. Se a
composição de amostra for significantemente diferente das amostras de água típicas, calibre o
amostrador com o líquido a ser coletado. Todos os programas devem ser parados antes da
calibração.
Calibre o sensor de líquido
A precisão do sensor de líquido varia com o tipo de líquido que está sendo amostrado. Por exemplo,
o volume de uma amostra altamente túrbida poderá não ser tão preciso como aquele de água limpa.
O sensor de líquido pode ser calibrado com água de amostra para aumentar a precisão de volume.
98
Português
1. Coloque a tubulação de entrada na água de amostra.
2. Pressione MENU.
3. Vá até PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuração do
programa>Modificar selecionado(s)>Sensor de líquido>Habilitar) para garantir que o sensor de
líquido esteja habilitado.
4. Acesse SYSTEM SETUP>LIQ SENSOR CAL (Configuração do sistema>Cal. do sensor de
líquido).
5. Selecione PERFORM CAL (Executar calibração).
6. Selecione START (Iniciar). A bomba opera em reverso para purgar o tubo de entrada. Em
seguida, a bomba opera na direção para frente.
7. Observe o fluxo da amostra no tubo de entrada para verificar se a amostra está passando pelo
sensor de líquido.
8. Quando a amostra estiver passando pelo sensor de líquido e não houver a presença de bolhas,
pressione STOP (Parar). O tubo de entrada será automaticamente purgado. A calibração sensor
de líquido está concluída.
9. Selecione DONE (Concluído) para sair e REPEAT (Repetir) para fazer a calibração novamente.
Como restaurar a calibração padrão
É possível voltar a calibração de volume do sensor de líquido para as configurações padrão de
fábrica.
1. Selecione PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuração
do programa>Modificar selecionado(s)>Sensor de líquido>Habilitar) para garantir que o sensor
de líquido esteja habilitado.
2. Pressione VOLUME CALIBRATION (Calibração de volume) e selecione RESET CAL (Redefinir
calibração). A calibração de volume será definida com as configurações padrão. Será mostrada a
mensagem “value has been reset to 0” (valor redefinido como 0).
Calibre o volume de amostra com o sensor de líquido
Faça a calibração do volume de amostra com o sensor de líquido habilitado para ajustar levemente
os volumes de amostra.
1. Vá até PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuração do
programa>Modificar selecionado(s)>Sensor de líquido>Habilitar) para garantir que o sensor de
líquido esteja habilitado.
2. Coloque a tubulação de entrada na água de amostra ou em água de torneira.
Observação: Use a água de amostra para obter uma precisão melhor.
3. Coloque tubulação da saída da bomba em um cilindro graduado.
4. Pressione VOLUME CALIBRATION (Calibração de volume) e selecione CALIBRATION
(Calibração).
5. Selecione START (Iniciar). A bomba opera em reverso para purgar o tubo de entrada. Em
seguida, a bomba opera na direção para frente e coleta o volume de amostra especificado no
programa de amostragem. A bomba opera em reverso para purgar o tubo de entrada.
6. Quando a amostra tiver sido coletada, selecione DONE (Concluído). Compare o volume da
amostra no cilindro com graduação com o volume de amostra que está inserido no menu de
configuração do programa.
7. Se o volume que foi coletado for diferente do volume de amostra no programa atual, insira o
volume que foi realmente coletado. Pressione ENTER. O sensor é ajustado para medir com
precisão todos os volumes programados.
8. Selecione START (Iniciar) para iniciar o programa de amostragem ou CANCEL (Cancelar) para
sair do menu de calibração do volume.
Português
99
Calibre o volume de amostra manualmente
Quando o sensor de líquido estiver desabilitado, todos os volumes deverão ser calibrados
manualmente por tempo. O volume de amostra está calibrado para o volume especificado no
programa atual. Se o volume de amostra for mudado no programa, o volume de amostra deverá ser
calibrado novamente para o novo volume.
1. Selecione PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>DISABLE
(Configuração do programa>Modificar selecionado(s)>Sensor de líquido>Desabilitar) para
garantir que o sensor de líquido esteja desabilitado.
2. Coloque a tubulação de entrada na água de amostra ou em água de torneira.
3. Coloque tubulação da saída da bomba em um cilindro graduado.
4. Pressione VOLUME CALIBRATION (Calibração de volume). Se for exibido mais de um volume,
selecione um dos volumes para a calibração.
5. Selecione START (Iniciar). A bomba opera em reverso para purgar o tubo de entrada. A bomba
então opera na direção para frente e começa a coletar amostras.
6. Pare a bomba no volume selecionado ou local de lavagem:
Volume de amostra: pressione STOP (Parar) quando o volume especificado no programa
principal for coletado.
Enxágue: pressione STOP (Parar) quando o líquido na tubulação de entrada alcançar o
detector de líquido.
Volume pluvial: pressione STOP (Parar) quando o volume especificado no programa águas
pluviais for coletado.
7. Selecione DONE (Concluído) para ir para a calibração do próximo volume ou REPEAT (Repetir)
para fazer a calibração novamente.
8. Quando todos os volumes tiverem sido calibrados, selecione DONE (Concluído).
9. Conecte a tubulação de saída ao encaixe de tubo no amostrador.
10. Selecione START (Iniciar) para iniciar o programa de amostragem ou CANCEL (Cancelar) para
sair.
Verifique o volume de amostra
Para garantir que o volume de amostra esteja preciso, colha uma amostra rápida. Não volte na
calibragem para verificar o volume já que a compensação de volume está reajustada em zero no
começo de uma calibragem.
1. Pressione MANUAL OPERATION (Operação manual).
2. Selecione GRAB SAMPLE (Amostra rápida).
3. Coloque a tubulação de entrada na água de amostra.
4. Coloque tubulação da saída da bomba em um cilindro graduado.
5. Se o sensor de líquido estiver habilitado, insira o volume a ser verificado.
6. Se o sensor de líquido estiver desabilitado, insira o volume que está especificado no programa.
7. Pressione ENTER. O ciclo da bomba é iniciado (purgação-amostra-purgação).
8. Compare o volume que foi coletado no cilindro graduado com o volume esperado. Se o volume
coletado não estiver correto, faça novamente a calibração de volume.
Manutenção
P E R I G O
Vários perigos. Somente pessoal qualificado deve realizar as tarefas descritas nesta seção do manual.
100 Português
P E R I G O
Risco de eletrocussão. Retire o instrumento da energia antes de realizar atividades de manutenção ou
serviço.
A D V E R T Ê N C I A
Exposição a riscos biológicos. Siga os protocolos de manuseio com segurança durante o contato com
os frascos de amostras e componentes do amostrador.
A D V E R T Ê N C I A
Vários perigos. O técnico deve garantir que o equipamento funcione segura e corretamente após os
procedimentos de manutenção.
Como limpar o instrumento
C U I D A D O
Perigo de incêndio. Não use agentes inflamáveis para limpar o instrumento.
A V I S O
Não limpe o aquecedor do compartimento do controlador com líquidos de qualquer natureza.
Se a água não for suficiente para limpar o controlador e a bomba, desconecte o controlador e mova-
o para longe do amostrador. Aguarde tempo suficiente para que o controlador e a bomba sequem
antes de reinstalar as peças e retomar os serviços.
Limpe o amostrador da seguinte maneira:
Gabinete do amostrador — limpe as superfícies internas e externas com um pano úmido e
detergente suave. Não use limpadores ou solventes abrasivos.
Garrafas de amostra — limpe as garrafas com uma água e escova e detergente suave. Enxágue
com água limpa e, em seguida, água destilada. Use uma autoclave para limpar as garrafas de
vidro, se necessário.
Substituir o dessecante
Um cartucho dessecante no controlador absorve a umidade e evita a corrosão. Monitore a cor do
dessecante através do visor (Figura 12). O dessecante em boas condições é amarelo. Quando sua
estiver verde, substitua o dessecante.
1. Desatarraxe e remova o cartucho dessecante (Figura 12).
2. Remova o plugue e descarte o dessecante gasto.
3. Encha o tubo dessecante com dessecante fresco.
4. Instale o plugue.
5. Aplique graxa ao O-ring.
6. Instale o tubo dessecante no controlador.
Português
101
Figura 12 Cartucho dessecante
1 Plugue 3 O-ring
2 Tubo de dessecante 4 Visor do dessecante
Manutenção da bomba
C U I D A D O
Risco de pressionamento. Desconecte o instrumento da alimentação antes de realizar atividades de
manutenção ou serviço.
Como substituir a tubulação da bomba
A V I S O
O uso de tubulação diferente daquela fornecida pelo fabricante poderá causar desgaste excessivo das peças
mecânicas e/ou desempenho ruim da bomba.
Examine a tubulação da bomba para ver se há desgaste no local em que os cilindro entram em
fricção com a tubulação. Substitua a tubulação quando a mensagem de lembrete for exibida ou
quando a tubulação mostrar sinais de desgaste.
Pré-requisitos:
Tubulação da bomba: pré-cortada ou longa de 4,6 m ou 15,2 m (15 pés ou 50 pés)
1. Desconecte a energia do controlador.
2. Se for usada a tubulação longa, corte a tubulação e adicione pontos de alinhamento. Consulte o
manual expandido no site do fabricante.
3. Remova a tubulação da bomba como mostrado nas etapas ilustradas que seguem.
4. Limpe o resíduo de silicone do interior do alojamento da bomba e dos cilindros.
5. Instale a nova a tubulação da bomba como mostrado nas etapas ilustradas que seguem.
102
Português
Português 103
Como limpar o rotor
Limpe o rotor, os trilhos do tubo da bomba e o alojamento da bomba com detergente suave.
Consulte Como substituir a tubulação da bomba na página 102 e as etapas ilustradas seguintes.
Como substituir o tubo do braço distribuidor
O braço distribuidor se move sobre cada garrafa durante amostragem de garrafas múltiplas.
Substitua o tubo do braço do distribuidor quanto o tubo estiver gasto. Certifique-se de que o tubo
correto é usado para o distribuidor correto e o braço distribuidor.
Observação: A tubulação do distribuidor não é a mesma que a tubulação da bomba. A tubulação da bomba
instalada no conjunto do distribuidor pode danificar o distribuidor. Além disso, as amostras podem ser perdidas,
pois o braço de distribuidor não se move com facilidade.
1. Remova o tubo do braço do distribuidor e do teto da do gabinete do amostrador.
2. Insira o novo tubo no braço do distribuidor. Certifique-se de que o tubo esteja alinhado com as
dimensões indicadas em Figura 13.
3. Insira a outra extremidade do tubo no encaixe do teto da do gabinete do amostrador.
4. Faça o teste de diagnóstico do distribuidor para garantir que está funcionando corretamente.
Figura 13 Conjunto do distribuidor
1 Bico 3 Motor distribuidor 5 Tubo do distribuidor
2 Braço distribuidor 4 Eixo
104 Português
Substituição de fusível
O fusível não é um item que pode ser reparado pelo usuário. Entre em contato com o suporte
técnico. Um fusível queimado pode ser uma indicação que o amostrador tem um problema e que o
serviço é necessário.
Português 105
目录
规格 106 操作 119
基本信息 107 维护 124
安装 110
附加信息
制造商网站中提供了附加信息
规格
产品规格如有变化,恕不另行通知。
SD900 便携式采样器
规格 标准底座 小型底座 复合底座
尺寸 直径:50.5 厘米(19.9 英寸)
高度:69.4 厘米(27.3 英寸)
直径:44.1 厘米(17.4 英寸)
高度:61 厘米(24 英寸)
直径:50.28 厘米
19.8 英寸)
高度:79.75 厘米
31.4 英寸)
重量 15 kg (35.6 lb),装 1-L 聚乙烯瓶
24 个)
14.8 kg (32.6 lb),装 10-L (2.5 gal) 聚乙
烯容器(1 个)
12.2 kg (27 lb),装 575-mL
(19.44 oz) 聚乙烯瓶(24 个)
12.9 kg (28.3 lb),装 10-L
(2.5 gal) 聚乙烯容器(1 个)
15 kg (36 lb),装 950-mL
(32.12 oz) 玻璃瓶
12 个)
外壳 耐冲击 ABS3 节构造;双壁底座,带 2.54 cm (1 in.)保温层 - 直身瓶,与冰直接接触。
样品温
0–60 °C (32–140 °F)
过滤器 标准尺寸、适合于较浅应用的高流速或低收缩 316 不锈钢、Teflon
®
或标准尺寸 316 不锈钢
样品抽
取试管
9.5 mm (3/8 in.)I.D. 乙烯基或特氟龙衬里聚乙烯
采样瓶
容量
1-L (0.26 gal) 聚乙烯和/ 350-mL
(11.83 oz) 玻璃瓶(24 个)
2.3-L (0.6 gal) 聚乙烯和/ 1.9-L
(0.5 gal) 玻璃瓶(8 个)
3.8-L (1 gal) 聚乙烯和/ 3.8-L (1 gal)
璃瓶(4 个)
3.8-L (1 gal) 聚乙烯和/ 3.8-L (1 gal)
璃瓶(2 个)
21-L (5.5 gal) 15-L (4 gal) 聚乙烯复合
容器或 20-L (5.25 gal) 聚乙烯或 10-L
(2.5 gal) 聚乙烯或 10-L (2.5 gal) 玻璃瓶
1 个)
575-mL (19.44 oz) 聚乙烯瓶
24 个)
950-mL (32.12 oz) 玻璃瓶
8 个)
10-L (2.5 gal) 聚乙烯瓶(1 个)
10-L (2.5 gal) 玻璃瓶(1 个)
21-L (5.5 gal) 聚乙烯瓶
1 个)
SD 控制器
规格 详细信息
尺寸 (W x H x D)
29.2 x 17.1 x 26.4 cm (11½ x 6¾ x 10
3
/
8
in.)
重量
4.2 kg (9.26 lb)
外壳 PC/ABS 复合材料,NEMA 4X6IP 67
106
中文
规格 详细信息
电源要求 由一个 8754500 电源供应 15 V 直流电;
过载保护 泵用 7 A 直流线路保险丝
操作温度 0 50 °C32 122 °F
存储温度 –30 60 °C (–22 140 °F)
存放/操作湿度 100% 冷凝
高速蠕动泵,带有弹簧安装式尼拉特隆滚柱
泵外壳
IP37
泵管 9.5 mm(内径)x 15.9 mm(外径)(
3
/
8
in.(内径)x
5
/
8
in.(外径))硅树脂管
泵管寿命 20000 次采样循环:1 L (0.3 gal) 采样量,1 次冲洗,6 分钟步调间隔,4.9 m (16 ft)
3
/
8
in.进样管,4.6 m (15 ft) 垂直升程,21 °C (70 °F) 采样温度
垂直采样升程 最大 8.5 m (28 ft)8.8 m (29 ft)
3
/
8
in.乙烯基进样管处于海平面高度且温度为 20–25 °C
(68–77 °F)
泵流速 4.8 L/min (1.25 gpm),其中垂直升程为 1 m (3 ft),并且采用
3
/
8
in.典型进样管
采样量 能以 10 mL (0.34 oz) 程控间隔从 10 mL 递增至 10000 mL ( 3.38 oz 递增到 2.6 gal)
样品量重复性(典型
值)
±5% × 200 mL 采样量,其中使用未经校准的液样检测方式:4.6 m (15 ft) 垂直升程,
4.9 m (16 ft)
3
/
8
in.乙烯基进样管,单瓶,满瓶关闭,室温,海拔 1524 m (5000 ft)
采样量精度(典型值) ±10% × 200 mL 进样量,其中使用未经校准的液样检测方式:4.6 m (15 ft)垂直升程,
4.9 m (16 ft)
3
/
8
in.乙烯基进样管,单瓶,满瓶关闭,室温,海拔 1524 m (5000 ft)
采样模式 步调:固定时间,固定流量,可变时间,可变流量,常数时间可变样品量 (CTVV)
请参阅本手册扩展版本中的常数时间可变样品量 (CTVV) 采样。
分布:单瓶复合、多瓶复合、多瓶分离、每个样品有多瓶、每瓶有多个样品、每个样品
有多瓶、每瓶有多个样品
运行模式 按照用户输入的采样次数连续运行或断续运行
多个程序 存储至多三个采样程序
输送速率(典型值) 0.9 m/s (2.9 ft/s)4.6 m (15 ft) 垂直升程,4.9 m (16 ft)
3
/
8
in.乙烯基进样管,温度
21 °C (70 °F),海拔 1524 m (5000 ft)
液样传感器 超声波式。主体:通过了 Ultem
®
NSF ANSI 标准 51,符合 USP Class VI
采样历史记录 最多 510 个记录
空气吹扫 每次采样前后自动进行空气吹扫。持续时间将会自动补偿进样管长度的变化。
事件日志 510 条记录
连接 电源连接、辅助连接、串行通信连接、分配器连接、SDI-12
过流材料 不锈钢、聚乙烯、聚四氟乙烯、聚醚酰亚胺、硅树脂
通信
RS232, Modbus, SDI-12
保修期 1
基本信息
对于因本手册中的任何不足或遗漏造成的直接、间接、特别、附带或结果性损失,制造商概不负责。
制造商保留随时更改本手册和手册中描述的产品的权利,如有更改恕不另行通知或承担有关责任。修
订版可在制造商的网站上找到。
中文
107
安全信息
对于误用和滥用造成的产品损坏,制造商概不负责,包括但不限于:直接、附带和间接的损坏,并且对于适用法律
允许的最大程度的损坏也不承担任何责任。用户唯一的责任是识别重大应用风险和安装适当的系统,以在设备可能
出现故障时保护工艺。
请在拆开本设备包装、安装或使用本设备前,完整阅读本手册。特别要注意所有的危险警告和注意事
项。否则,可能会对操作者造成严重的人身伤害,或者对设备造成损坏。
确保设备提供的保护没有受损。请勿以本手册指定方式之外的其它方式使用或安装本设备。
危险信息使用
表示潜在的或紧急的危险情况,如果不加以避免,将会导致死亡或严重伤害。
表示潜在或非常危险的情形,如不避免,可能导致严重的人身伤亡。
表示潜在的危险情形,可能导致一定程度的人身伤害。
表明如不加以避免则会导致仪器损坏的情况。需要特别强调的信息。
警告标签
请阅读贴在仪器上的所有标签和标记。 如未遵照这些安全标签的指示操作,则可能造成人身伤害或仪
器损坏。 仪器上的符号在手册中通过警告说明参考。
这是安全警报标志。请遵守此标志后面的所有安全信息,以避免可能造成的伤害。如果仪器上有此标
志,则请参见仪器手册,了解操作或安全信息。
此标记指示存在潜在的夹伤危险。
标有此符号的电气设备在欧洲不能通过家庭或公共垃圾系统进行处理。请将老旧或报废设备寄回至制造
商处进行处置,用户无需承担费用。
认证
加拿大无线电干扰产生设备法规(Canadian Radio Interference-Causing Equipment
Regulation),IECS-003A
制造商支持测试记录留存。
A 类数字设备符合加拿大干扰产生设备法规的所有要求。
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC 15 部分,“A”类限制
制造商支持测试记录留存。该设备符合 FCC 规定第 15 部分的要求。设备操作满足以下两个条件:
1. 本设备不会造成有害干扰。
2. 本设备必须接受任何接收到的干扰,包括可能导致意外操作的干扰。
108
中文
若未经负责出具符合声明的一方明确同意擅自对本设备进行改动或改装,可能会导致取消用户操作该
设备的权限。本设备已经过测试,符合 FCC 规定第 15 部分中确定的 A 类数字设备限制。这些限制专
门提供当设备在商业环境下工作时针对有害干扰的合理保护。该设备产生、使用和放射无线电射频能
量,如果不按照说明手册的要求对其进行安装和使用,可能会对无线电通讯造成有害干扰。本设备在
居民区工作时可能会产生有害干扰,这种情况下用户须自行承担费用消除这种干扰。以下方法可用于
减少干扰问题:
1. 断开设备的电源,以便确证它是干扰源与否。
2. 如果设备与遭受干扰的仪器连接到相同的插座,将设备连接到其他插座。
3. 将设备从接受干扰的仪器边上移开。
4. 重新定位受干扰仪器的接收天线。
5. 同时尝试以上多项措施。
产品概述
化学或生物危害。如果该仪器用于监测处理过程和/或监测有法规限制以及有与公众健康、公众安全、食
品或饮料制造或加工相关监测要求的化学药物添加系统时,仪器的使用者有责任了解并遵守所有适用的
法规,并且要建立适当的机制,确保在仪器发生故障的时候也不会违法这些法规。
火灾危险。此产品不得与易燃液体一同使用。
SD900 便携式采样器按指定的间隔时间采集液体样本,并将样品存放在采样瓶或容器中。该采样器适
用于各种水体应用以及有毒污染物和悬浮物。安装采样器和各种固定座、采样瓶或容器。请参阅
1
1 产品概述
1 小型底座 6 控制器接线
2 标准绝缘底座 7
3 中心部分 8 控制器
4 电源 9 液样传感器
5 顶盖 10 标准绝缘底座,适合 21-L (5.5 gal) 容器
产品组件
确保已收到所有组件。 请参阅 2 如有任何物品丢失或损坏,请立即联系制造商或销售代表。
中文
109
2 产品组件
1 底座(标准型、小型或复合型) 6 电池充电器(选件)
2 单瓶安装的组件(瓶子和支架可更换) 7 交流电源(可选)
3 中心部分(带控制器) 8 过滤器
4 电池(选件) 9 进样管,乙烯基或特氟龙衬里
5 顶盖 10 多瓶安装的组件(瓶子和固定座可更换)
安装
多种危险。只有合规的专业人员才能从事文档中本部分所述的任务。
密闭空间预防措施
爆炸危险。在进入密闭空间之前,需要进行有关进入之前测试、通风、进入过程、疏散救援过程以及安
全工作做法的培训。
下列信息旨在帮助用户了解进入密闭空间时可能遇到的危险和风险。
1993 4 15 日,OSHA 最终颁布法律规定 CFR 1910.146,密闭空间作业许可 (Permit Required
Confined Spaces)。该标准对美国 250000 多个工业场地具有直接影响,用于保护狭窄空间内工人的
健康和安全。
密闭空间定义:
密闭空间是指具有下列一种或多种情况(或具有直接可能性)的任何位置或壳体:
氧气浓度低于 19.5% 或高于 23.5% 并且/或者硫化氢 (H
2
S) 浓度高于 10 ppm 的空气。
含有易燃易爆的气体、蒸气、雾气、粉尘或纤维的空气。
有毒材料,接触或吸入时会引起受伤、损害健康或致死。
密闭空间不适用于人类居住。密闭空间限制进入,其中含有已知的或潜在的危害。密闭空间包括检修
孔、堆栈、管道、大桶、开关地下室及其他类似位置。
进入可能含有危险气体、蒸气、雾气、粉尘或纤维的密闭空间和/或位置之前,务必遵守标准的安全规
程。进入密闭空间之前,请首先查找并阅读所有相关的规程。
110
中文
机械安装
现场安装指导
爆炸危险。该仪器未获批准在危险场所安装。
请参阅现场位置评估指南。
如果采样器的安装空间有限,请遵循所有安全保护措施。请参阅 密闭空间预防措施 110
确保该位置的温度在规范范围内。请参阅 规格 106
将采样器安装在平面上或使用悬架装置、支架或拉紧杆悬挂采样器。请参阅 将采样器安装在检修孔
111 和适用的安装文档。
尽可能接近样品源以降低分析延迟。请参阅 使采样器保持垂直 114
有关载运速度和最大垂直扬程的限制信息,请参阅 规格 106
将采样器安装在检修孔中
将采样器安装在检修孔中样水上方。为采样器安装拉紧杆或支架。将拉紧杆安装在检修孔内部。拉紧
杆受压支固在侧壁上。支架的宽度与检修孔护盖宽度相同。在支架护盖的正下方安装支架。请参阅制
造商网站上提供扩展版手册。请参阅随附件提供的文档安装采样器。
安装分配器或满瓶关断装置(选件)
分配器或满瓶关断装置组件一般出厂时已安装。请参阅分配器或满瓶关断装置文档了解安装说明。
准备采样器
清洁采样瓶
用刷子、水及温和的清洁剂清洁采样瓶和瓶盖。先后用淡水及蒸馏水冲洗容器。
安装一个采样瓶
利用一个采样瓶采集一种混合样品。采样瓶满样时,满瓶关闭功能将会停止采样程序。
1. 清洁采样瓶。请参阅 清洁采样瓶 111
2. 按照 3 所示安装采样瓶。
中文
111
3 单瓶安装
1 聚乙烯瓶,10 L (2.5 gal) (1918) 6 聚乙烯瓶,21 L (5.5 gal) (6494)
2 玻璃瓶,10 L (2.5 gal) (6559) 7 小型底座 (8975)
3 支架 (1502) 8 标准绝缘底座 (8976)
4 聚乙烯瓶,15 L (4 gal) (1367) 9 标准绝缘底座,适合 21 L (5.5 gal) 的采样瓶 (8561)
5 聚乙烯瓶,19 L (5 gal) (6498)
为多个采样安装第一个采样瓶
用多个采集瓶采集样品,装入单独的采集瓶或多个采集瓶。分配器将采样管移至每个采样瓶上方。按
4 中所示将采样瓶安装在采样器底座中。将第一个采样瓶(1 号)安装在采样器底座标签下方。
根据标签指示的方向,按升序逐号安装其余的采样瓶。请参阅 5 114 查看必要的组件图。
112
中文
4 1 号瓶安装
1 1 号瓶安装位置(小型底座) 6 弹力带
2 1 号瓶位置,24 个瓶 7 标准绝缘底座,适合 21 L (5.5 gal) 的采样瓶
3 1 号瓶位置,8 个瓶 8 标准绝缘底座
4 1 号瓶位置,2 个或 4 个瓶 9 小型底座
5 固定座
安装多个采样瓶
安装多个采样瓶时,分配臂会使采样管移动到每个瓶子的上方。采集的样品达到指定数量时,自动结
束采样。
1. 清洁采样瓶。请参阅 清洁采样瓶 111
2. 在采样器底座标签下方安装第一个采样瓶(1 号)请参阅 为多个采样安装第一个采样瓶 112
3. 按照 5 所示装配采样瓶。对于 8 个或更多采样瓶,请确保第一个采样瓶以顺时针方向紧靠第一
个瓶符进行放置。
4. 将采样瓶组件放在采样器中。对于 8 个或更多采样瓶,对齐底部瓶托槽中的接线。
中文
113
5 多瓶安装
1 定位座/固定座 (2347) 8 玻璃瓶,3.8 L (1 gal)4 个)(2216),或聚乙烯瓶,
3.8 L (1 gal)4 个)(2217)
2 固定座 (1422) 9 玻璃瓶,1.9 L (0.5 gal)8 个)(1118),或聚乙烯
瓶,2.3 L (0.61 gal),8 个)(657)
3 固定座 (2190) 10 聚乙烯瓶,1 L (0.26 gal)24 个)(737)
4 固定座 (2189) 11 玻璃瓶,350 mL (11.83 oz)24 个)(732)
5 玻璃瓶,950 mL (32.12 oz)8 个)(2348) 12 小型底座 (8975)
6 聚乙烯瓶,575 mL (19.44 oz)24 个)(1369) 13 标准底座 (8976)
7 玻璃瓶,3.8 L (1 gal)2 个)(2214),或聚乙烯瓶,
3.8 L (1 gal)2 个)(2215)
使采样器保持垂直
在采样流中间部位安装过滤器(不要在表面或底部安装),确保采集到具有代表性的样品。请参阅
6 请参阅 7 了解进样管的安装。
1. 按照 7 所示将进样管连接到采样器。
注:
如果使用特氟龙衬里的管子,则采用连接套件
2186
2. 将进样管和过滤器安装在样品源的主液流中,即水处于湍流状态并且混合均匀的位置。
使进样管尽可能短。
使进样管保持最大的垂直斜度,以便在两次采样期间完全排空管子。
注:
如果无法保持垂直倾斜,或是管子加压,则禁用液体传感器。手动校准采样量
.
请参阅
手动校准采样量
124
确保进样管未扭曲。
114
中文
6 仪器安装
1 过滤器 3 进样管
2 垂直上升
7 进样管安装
电气安装
控制器接线
电击危险。外部连接的设备必须具有适用的国家安全标准评估。
8 所示是控制器的接线。
中文
115
8 控制器接头
1 辅助装置 4 SDI-12 装置选项
2 电源 5 分配器/满瓶关闭
3 串行通信
连接一个流量计
将一个流量计连接到控制器,以便在样品流量高于或低于指定值时启动或关闭采样器。
需收集的物品:
Sigma 流量计用多功能全电缆(或 980 型流量计用 980 半电缆)。
可选分接头增加连接。两个或多个分接头可以串行连接。
1. 将电缆一端连接至流量计。对于 980 型流量计,请参阅 980 型流量计用户手册。
2. 将电缆的另一端连接到控制器的辅助装置接头上。
注:
如果流量计采用一根
6
引脚电缆,则使用
6
引脚转
7
引脚适配电缆。
连接一台个人电脑或通信网络
将一台个人电脑或通信网络连接到控制器,以便传送数据或配置采样器。
需收集的物品:
串行电缆,7 引脚,由 RS232 接口转接到 DB-9 接口(请参阅 9)。 接线:B 5(信号接
地);D 3 (RCD)F 2 (TXD)G(接地)。
1. 将串行电缆的一端连接到通信设备或网络上。
2. 将电缆的另一端连接到控制器的串行通信接头上。
9 串行通信电缆
将采样器连接至电源
电击致命危险。如果此设备在户外或在可能潮湿的场所使用,则必须使用接地故障电路中断器
GFCI/GFI 将此设备连接到其主电源。
116
中文
火灾危险。在电源线中安装 15 A 断路器。断路器如果极为贴近设备,则可以是本地电源断开装置。
电击致命危险。需要连接保护接地地线。
电击致命危险。确保可轻松切断本地电源。
电击危险。如果泵循环之间的时间太短,则电源可能温度过高。
确保泵的循环时间不会造成电源过热。请参阅 计算泵的循环时间 117
使用电池或交流电源为采样器供电。请参阅下面所示的说明步骤。
计算泵的循环时间
在两次采样循环中间,泵必须关闭至少一段时间,以便电源不会过热。
1. 计算泵在采样循环期间连续运行的总时间。包括所有阶段:清洗前、吸入清洗、采样、采样重试和
清洗后。
2. 根据 10 确定两次采样循环期间泵保持关闭的最短时间。
3. 确保采样器程序的步调或时间间隔让泵在两次采样循环中间关闭至少一段时间。
中文
117
10 泵在 50 °C (122 ºF) 下启动和关闭的时间
1 泵必须关闭的时间(以分为单位) 3 示例:如果泵连续运行 5 分钟,则泵必须关闭 15
钟。
2 泵连续运行的时间(以分为单位)
开启电源
按下电源键接通或断开电源。确保电源(交流电源或电池供电)正确安装。
用户界面及导航
控制器键盘如 11 所示。利用 MENU(菜单键) 可以设置采样程序、配置控制器的设置或完成诊断
测试。利用箭头键、ENTER(输入键)及 BACK(返回键)可以滚动浏览菜单、进行选择及输入数
值。查看显示屏幕底部或边侧的箭头可以了解更多可用选项。
118
中文
11 控制器键盘
1 POWER(电源) 5 MENU(菜单) 9 STATUS(状态)
2 VOLUME CALIBRATION(采样
量校准)
6 ENTER(输入) 10 RUN/HALT PROGRAM(运
/中止程序)
3 STOP(停止) 7 BACK(返回) 11 MANUAL OPERATION(手动
操作)
4 LEDLED 指示灯) 8 ARROW KEYS(箭头键)
操作
配置系统的设置
1. 按下 MENU(菜单)
2. 进入 SYSTEM SETUP(系统设置)。
3. 更改可用选项。
选项 说明
Time/date setup(时间/日期设
置)
设置时间(24 小时格式) 和日期。
Communication(通信) 为串行端口选择波特率(192003840057600 115200)和协议
(Modbus RTU ASCII)。
Setup base(底座设置) 选择采样器底座(全天候冷藏式、冷藏式、便携式标准型或便携式紧凑
型)。
Language(语言) 选择控制器的语言。
Liq sensor cal(液样传感器校
准)
校准液样传感器或设为默认。
Set contrast(设置对比度) 调整 LCD 屏幕的对比度。
Pump duty cycle(泵的负载周
期)
更改泵的负载周期。周期范围:50%100%(默认:100%)。
中文
119
选项 说明
Password setup(密码设置) 设置一个密码,以控制对程序设置菜单及系统设置菜单的访问。选择
ENABLE(启用)>YES(是)以设置一个新密码,或选择 ENABLE(启
用)>NO(否)以激活当前密码。如果首次设置密码,则使用 900900
为当前密码。
Tubing life(管子寿命) 设置一个泵管更换提示。进入 ENABLE(启用)>RESET CYCLES(重置
循环),将泵循环计数重置为 0。进入 ENABLE(启用)>CYCLE LIMIT
(循环限制)以更改所提示的泵循环数。进入 ENABLE(启用)
>STATUS(状态)以显示当前的泵循环计数。
SDI–12 (optional)SDI–12(可
选))
配置 SDI–12 装置的运行。只有在发现一个装置时,才会显示该选项。
设置一个采样程序
采样程序包括自动采样所需的所有参数。定期采样或在采样程序结束时进行采样。
1. 按下 MENU(菜单)
2. 进入 PROGRAM SETUP(程序设置)>MODIFY ALL(更改所有项)。然后会显示第一个屏幕。
3. 为每个参数选择或输入数值。利用左右箭头键可左右移动。利用上下箭头键可更改数值。如有必
要,用箭头键修改单位(gal mL)。
选项 说明
Bottles(采样
瓶)
采样瓶数量:采样器中的瓶子数量(1248 24)。
满瓶关闭:启用/禁用。如果启用该功能,则控制器会检查采样瓶是否出现溢流状况。
瓶容量:每个采样瓶的体积容量。容量范围:50–65000 mL (0.01 - 17.17 gal)
Intake tubing(进
样管)
管长:过滤器至液样传感器之间的进样管长度。长度范围:100–3000 cm 3–99 ft
确的管长才能获得精确的采样量。
进样管类型:进样管的直径和材料(¼ in. 乙烯基塑料管、3/8 in. 乙烯基塑料管或 3/8 in.
特氟龙管)。
Program delay
(程序延迟)
启用/禁用:选择启用时,采样程序会在指定时刻或计数下启动。日期和时间:程序启动
时的日期和时间(24 小时格式)。计数:程序将要启动时,流量计的计数或脉冲数(1
9999 个计数)。如果采样步调随后设为时间形式,则程序延迟功能将被禁用。
Sample pacing/
collection(采样
步调/样品采集)
基于时间 - 请参阅 基于时间的采集 121
注:
当使用
CTVV
时:
始终启用液体传感器。
样品分布为复合式。
运行模式为非持续性。
间隔时间之后采集第一个样品。
流量源通常为
4–20 mA
输入。
基于流量 - 请参阅 基于流量的采集 121
Sample
distribution(样
品分布)
Deliver samples to all(将样品输送到所有采样瓶)>YES(是)
每个样品均被输送到所有采样瓶。采样结束:最后一次采样之后结束程序 (1–999)。连续
采样:继续执行采样程序,直到手动停止为止。
Deliver samples to all(将样品输送到所有采样瓶)>NO(否)
将样品输送到一部分采样瓶中。样品/采样瓶:每个采样瓶需要采集的样品数量 (1–999)
采样瓶/样品:含有相同样品的采样瓶的数量。
Liquid sensor
(液样传感器)
启用或禁用。如禁用,采样量必须按时间校准。
Sample volume
(采样量)
每个样品的采集量,单位为 mL (1 gal = 3785.4 mL)。如果使用采样瓶/样品模式,则每个
采样瓶将会采集满瓶样品 (10–10000 mL)。确保采样量不超过采集瓶体积。采样量以
10 mL 为基础四舍五入。
120
中文
选项 说明
Intake rinses(进
样冲洗)
采样之前冲洗的进样管数量 (0–3)
Sample retries
(采样重试次数)
采样失败之后尝试的采样次数 (0–3)
Site ID(采样点
ID
采样位置的名称(最长 12 个字符)。保存程序时用采样点 ID 作为预设文件名。
Advanced
sampling(高级
采样)
进入高级采样菜单。必要时在 Advanced Sampling(高级采样)菜单中配置相应选项。
运行程序 基本的采样程序创建已完成。采样器将提示用户启动或取消采样程序。
基于时间的采集
1. 选择选项或为每个选项输入数值。
选项 说明
步调间隔 定时采样的时间间隔,以小时和分钟为单位 (0:01–999:00)
采集第一个样品 程序立即启动,或在第一个间隔时间之后启动。
可变样品量 指定样品量是否可变(基于流量)。如果 yes(是),请参阅 使用可变样品量 121。如果
no(否),请参阅 切勿使用可变样品量 121
使用可变样品量
样品步调为常数时间可变样品量 (CTVV)
1. 选择选项或为每个选项输入数值。
选项 说明
选择单位 选择要使用的流量单位。
设定 4mA 输入 输入与 4mA 输入相一致的流量。
设定 20mA 输入 输入与 20mA 输入相一致的流量。
平均流量 指定该位置的平均流量 (1–999,999)。这些单位由以上选定的单位指定。
所需的时间量 指定执行样品程序期间要采集的总量 (10–10,000 ml)
采集时长 指定总采集时长,以小时和分钟为单位 (0:01–999:00)
最小样本量 如果流量过低,不足以达到最小样本量,则跳过此样品 (10–10,000 ml)
切勿使用可变样品量
1. 为该选项选择或输入数值。
选项 说明
采集第一个样品 选择立即启动程序或在首次时间间隔过后再启动。
基于流量的采集
为基于流量的采样指定流量源:
计数 - 请参阅 针对流量源使用计数 122
4–20 mA 输入 - 请参阅 为流量源使用 4–20 mA 输入 122
中文
121
针对流量源使用计数
1. 选择选项或为每个选项输入数值。
选项 说明
每次达到计数时采集样品 以固定的流量间隔采样,以计数为单位(1–9999 个计数)。
超时时间 如果流量很低,则采集样品。
时间 两次采样之间的最大间隔时间 (0:01–999:00)。每次采样之后,计时器都会重新启
动。
采集第一个样品 程序立即启动,或在第一个间隔时间之后启动。
为流量源使用 4–20 mA 输入
1. 选择选项或为每个选项输入数值。
选项 说明
选择单位 选择要使用的流量单位。
设定 4mA 输入 输入与 4mA 输入相一致的流量。
设定 20mA 输入 输入与 20mA 输入相一致的流量。
每次达到计数时采集样
输入抽样之前的总流量。
超时时间 在流量异常低时,选择启用该功能进行采样。
时间 两次采样之间的最大间隔时间 (0:01–999:00)。每次采样之后,计时器都会重新启
动。
采样量 输入每个样品的采集量。如选择采集瓶/样品模式,每只采集瓶内一种样品的采样量
(100–10,000 mL)。如果选择了可变样品量,则将跳过此选项。
恢复默认设置
恢复选项用于将所有的程序参数设为默认值。
1. 按下 MENU(菜单)
2. 进入 PROGRAM SETUP(程序设置)>RESTORE(恢复)。
3. 选择 YES(是)。
启动或停止一个采样程序
利用 RUN/HALT PROGRAM(运行/中止程序)键可以启动或停止一个程序。可以设置多达 12 个用户
自定义的启动/停止时间/日期。更改程序或系统的设置之前,必须停止采样程序。
1. 按下 RUN/HALT PROGRAM(运行/中止程序)
2. 选择一个选项。
选项 说明
Start(启动) 启动当前加载的程序。其状态将变为 RUNNING(正在运行)。
Halt(中止) 临时停止程序。其状态将变为 HALTED(已中止)。
Resume(恢复) 该程序将从中止点恢复运行。其状态将变为 RUNNING(正在运行)。
Start over(重新开始) 采样程序将从开始点启动。其状态将变为 RUNNING(正在运行)。
End program(结束程序) 结束当前正在运行的程序。其状态将变为 COMPLETE(已结束)。
122
中文
手动操作
使用手动操作功能采集抓取样品、移动分配臂或操作泵。详情请参阅该手册扩充版。
采样量校准
采样量可以通过液样传感器自动进行校准,或采用手动校准。使用液样传感器校准体积时,调整传感
器,精确测量所有程式化体积。如果禁用了液样传感器,则必须手动校准采样量、冲洗量及雨水量。
对于典型应用,可以利用自来水进行采样量校准。如果样品的组成明显不同于典型的水样,则应利用
所需采集的液体来校准采样器。校准之前必须停止所有程序。
校准液样传感器
液体传感器的精度随采样液体的类型而异。例如,严重浑浊样品的采样量不会与清澈样品的采样量一
样精确。液体传感器可以利用水样进行校准,以提高采样量的精度。
1. 将进样管放入水样中。
2. 按下 MENU(菜单)
3. 进入 PROGRAM SETUP(程序设置)>MODIFY SELECTED(修改选择项)>LIQUID SENSOR
(液体传感器)>ENABLE(启用),以确保启用液体传感器。
4. 进入 SYSTEM SETUP(系统设置)>LIQ SENSOR CAL(液体传感器校准)。
5. 选择 PERFORM CAL(进行校准)。
6. 选择 START(开始)。泵反向运行以清洗进样管。随后泵以前进方向运行。
7. 请查看进样管中的样品流,确保样品流经液体传感器。
8. 样品流经液体传感器并且无气泡时,按下 STOP(停止)。进样管会自动进行清洗。液体传感器结
束校准。
9. 选择 DONE(结束校准)将会退出,或选择 REPEAT(重复校准)再次进行校准。
恢复默认校准
液样传感器的采样量校准可以恢复为默认出厂设置。
1. 选择 PROGRAM SETUP(程序设置)>MODIFY SELECTED(更改所选项)>LIQUID SENSOR
(液样传感器)>ENABLE(启用),以确保启用液样传感器。
2. 按下 VOLUME CALIBRATION(采样量校准),然后选择 RESET CAL(重置校准)。此时采样
量校准被设为默认设置。并会显示数值已被重置为 0”信息。
使用液体传感器校准采样量
在启用液体传感器的情况下校准采样量,以便微调采样量。
1. 进入 PROGRAM SETUP(程序设置)>MODIFY SELECTED(修改选择项)>LIQUID SENSOR
(液体传感器)>ENABLE(启用),以确保启用液体传感器。
2. 将进样管放入水样或自来水中。
注:
利用水样获得最佳精度。
3. 将泵出口侧的管子放入一个量筒中。
4. 按下 VOLUME CALIBRATION(采样量校准)并选择 CALIBRATION(校准)。
5. 选择 START(开始)。泵反向运行以清洗进样管。随后泵以前进方向运行,并采集采样程序中指
定的采样量。泵反向运行以清洗进样管。
6. 结束采样时,选择 DONE(结束)。将量筒中的采样量与程序设置菜单中输入的采样量进行对比。
7. 如采集体积与当前程序的采样量不同,输入实际采集的体积。按 Enter(输入)。传感器经过了调
整,以便精确地测量所有设置的采样量。
8. 选择 START(开始)以启动采样程序,或选择 CANCEL(取消)以退出采样量校准菜单。
中文
123
手动校准采样量
禁用液体传感器时,必须按时手动校准所有的采样量。采样量校准为当前程序规定的量。如果改变了
程序中的采样量,必须为新的采样量进行校准。
1. 选择 PROGRAM SETUP(程序设置)>MODIFY SELECTED(修改选择项)>LIQUID SENSOR
(液体传感器)>DISABLE(禁用),以确保禁用液体传感器。
2. 将进样管放入水样或自来水中。
3. 将泵出口侧的管子放入一个量筒中。
4. 按下 VOLUME CALIBRATION(采样量校准)。如果显示多个采样量,则选择其中一个进行校
准。
5. 选择 START(开始)。泵反向运行以清洗进样管。随后泵以前进方向运行,并开始采集样品。
6. 到达所选体积或冲洗位置时,停止泵运行:
采样量:获得主程序中指定的采样量时,按下 STOP(停止)
冲洗:进样管中的液体到达液体检测器时,按下 STOP(停止)
雨水采样量:获得雨水程序中指定的采样量时,按下 STOP(停止)
7. 选择 DONE(结束校准) 进入下一个采样量校准,或是选择 REPEAT(重复校准)再次进行校
准。
8. 校准完所有采样量时,选择 DONE(结束校准)。
9. 将出样管连接到采样器的管接头上。
10. 选择 START(开始)启动采样程序,或选择 CANCEL(取消) 退出采样。
验证采样量
为确保采样量准确,需要进行随机采样。不要返回校准菜单检查体积,因为在校准开始时体积补偿已
被重置为 O
1. 按下 MANUAL OPERATION(手动操作)
2. 选择 GRAB SAMPLE(随机采样)。
3. 将进样管放入水样中。
4. 将泵出口侧的管子放入一个量筒中。
5. 如果启用了液样传感器,则输入需要验证的采样量。
6. 如果禁用了液样传感器,则输入程序中指定的采样量。
7. 按下 Enter(输入)。随后泵循环将会启动(吹扫 - 采样 - 吹扫)。
8. 将量筒中采集的样品量与预期样品量进行对比。如果采集的样品量不正确,则重新进行采样量校
准。
维护
多种危险。只有合规的专业人员才能从事文档中本部分所述的任务。
电击致命危险。执行维护或维修活动前,请断开仪器的电源连接。
生物危害暴露危险。接触取样瓶和采样器部件时,请遵守安全处置规定。
124
中文
多种危险。技术人员必须确保完成维护程序后,安全正确地操作设备。
清洁仪器
火灾危险。切勿使用易燃试剂来清洁仪器。
请勿使用任何液体清洁控制器室加热器。
如果水量不足以清洁控制器和泵,则断开控制器,然后使控制器远离采样器。在重新安装零件并恢复
使用之前,让控制器和泵充分干燥。
按如下方式清洁采样器:
采样器外壳 - 使用湿布和温和的清洁剂清洁采样器外壳的内外表面。不得使用磨蚀性洗涤剂或溶
剂。
采样瓶 - 使用软刷、水和温和的清洁剂清洗取样瓶。先用清水冲洗,然后用蒸馏水冲洗。如有需
要,使用高压灭菌器清洁玻璃瓶。
更换干燥剂
控制器中的干燥剂筒用于吸收湿气,防止发生腐蚀。通过窗口 ( 12) 来监控干燥剂的颜色。新鲜的干
燥剂为黄色。颜色变绿时,需要更换干燥剂。
1. 拧松后拆下干燥剂筒 ( 12)
2. 取下塞子,丢弃旧干燥剂。
3. 将新鲜的干燥剂充满干燥剂管。
4. 安装塞子。
5. O 形环涂上密封脂。
6. 将干燥剂管安装在控制器中。
12 干燥剂筒
1 3 O 形环
2 干燥剂管 4 干燥剂窗口
中文
125
泵维护
夹伤危险。执行维护或维修活动前,请断开仪器的电源连接。
更换泵管
除了制造商供应的管子之外,使用其他的管子会造成机械零件过度磨损和/或使泵性能不良。
检查泵管是否因为滚柱摩擦管子而产生磨损。显示提示信息时,或是管子出现磨损迹象时,请更换泵
管。
处理前必要条件:
泵管 — 4.6 m 15.2 m15 ft 50 ft)预切管材或成批管材
1. 断开控制器的电源。
2. 如果使用成批管材,需要切割管材并标注对齐点。 参阅制造商网站上的扩展手册。
3. 按照随后的图示步骤拆下泵管。
4. 清洁泵壳内部和滚柱残留的硅树脂。
5. 按照随后的图示安装新泵管。
126
中文
清洁转子
用柔性洗涤剂清洁转子、泵管路及泵外壳。请参阅 更换泵管 126 及随后图示步骤。
中文
127
更换分配臂管
在多瓶采样过程中,分配器分配臂会移动到每个采集瓶上。分配臂中的管子磨损时,需要进行更换。
确保在分配器和分配器分配臂中使用正确的管。
注:
分配器导管系统不同于泵的导管系统。装在分配器组件内的泵导管可能损坏分配器。而且,样品会因为分配臂
无法轻易移动而漏失。
1. 从分配臂和中心部分顶部拆下管子。
2. 将新管子插入分配臂中。确保导管符合 13 中规定的尺寸。
3. 将管子的另一端插入中心部分顶部的接头中。
4. 对分配器完成诊断测试,以确保运行正常。
13 分配器组件
1 管嘴 3 分配器电机 5 分配器管
2 分配器分配臂 4 泵轴
保险丝更换
保险丝不能由用户更换。请联系技术支持部门。保险丝熔断则意味着采样器出现问题,需要维修。
128
中文
目次
仕様 ページの 129 操作 ページの 142
総合情報 ページの 131 メンテナンス ページの 148
設置 ページの 134
より詳しい情報
より詳しい情報は、当社 Web サイトにて入手できます。
仕様
この仕様は予告なく変更されることがあります。
SD900 Portable Sampler
仕様 標準ベース コンパクトベース 混合ベース
寸法 直径: 50.5 cm
高さ: 69.4 cm
直径: 44.1 cm
高さ: 61 cm
直径: 50.28 cm
高さ: 79.75 cm
重量 15 kg1 L ポリエチレンボトル (24x) を使
14.8 kg 10 L ポリエチレンコンテナー
(1x) を使用
12.2 kg575 mL ポリエチレン
ボトル (24x) を使用
12.9 kg10 L ポリエチレンコン
テナー (1x) を使用
15 kg950 mL ガラスボト
(12x) を使用
筐体 衝撃耐性のある ABS3 セクション構造、2.54 cm の二重壁ベース絶縁ボトルが氷と直に接触
試料
温度
0 60 °C
スト
レー
ナー
316 ステンレススチール (標準サイズ、高速、浅い水深用ロープロファイル、Teflon
®
加工) または
316 ステンレススチール (標準サイズ)
試料
取入
9.5 mm 内径、ビニールまたはテフロン加工ポリエチレン
試料
ボト
ルの
容量
1 L ポリエチレンおよび/または 350 mL
ガラスボトル (24x)
2.3 L ポリエチレンおよび/または 1.9 L
ガラスボトル (8x)
3.8 L ポリエチレンおよび/または 3.8 L
ガラスボトル (4x)
3.8 L ポリエチレンおよび/または 3.8 L
ガラスボトル (2x)
21 L もしくは 15 L ポリエチレン複合材
コンテナー、または 20 L ポリエチレン、
10 L ポリエチレン、もしくは 10 L ガラス
(1x)
575 mL ポリエチレンボトル
(24x)
950 mL ガラスボトル (8x)
10 L ポリエチレンボトル (1x)
10 L ガラスボトル (1x)
21 L ポリエチレンボトル
(1x)
SD コントローラー
仕様 詳細
寸法 (W x H x D)
29.2 x 17.1 x 26.4 cm
重量
4.2 kg
日本語
129
仕様 詳細
筐体 PC/ABS ブレンド、NEMA 4X6IP 67
電源 8754500 電源から 15 VDC が供給。 内蔵電源から 15 VDC が供給
過負荷保護 ポンプ用の 7 ADC 回路のヒューズ
動作周囲温度 0 50 °C
保管温度 –30 60 °C
保管/作動湿度 100 %、結露可
ポンプ スプリング式ナイラトロンローラーによる高速蠕動ポンプ
ポンプ筐体
IP37
ポンプチューブ 内径 9.5 mm x 外径 15.9 mm (内径
3
/
8
インチ x 外径
5
/
8
インチ) シリコン
ポンプチューブの寿命 20,000 サンプルサイクル: サンプル量 1 L、洗浄回数 1 回、ペース間隔 6 分、長さ
4.9 m
3
/
8
インチ取入管、揚水高さ 4.6 m、サンプル温度 21 °C の場合
サンプル揚水高さ 最大 8.5 m: 長さ 8.8 m
3
/
8
インチビニール製取入管、海抜、温度 20 25 °C の場
ポンプ流量 4.8 L/min (1.25 gpm): 揚水高さ 1 m
3
/
8
インチ取入管の場合
サンプル量 10 10,000 mL の範囲で 10 mL 単位で設定可能
サンプル量の繰り返し精
(標準値
サンプル量 200 mL ±5 %: 未校正の液体、揚水高さ 4.6 m、長さ 4.9 m
3
/
8
イン
チビニール製取入管、単一ボトル、フルボトルシャットオフ、室温、高度 1524 m
場合
サンプル量の精度 (標準
)
サンプル量 200 mL ±10 %: 未校正の液体、揚水高さ 4.6 m長さ 4.9 m
3
/
8
イン
チビニール製取入管、単一ボトル、フルボトルシャットオフ、室温、高度 1524 m
場合
サンプリングモード ペース: 時間固定、流量固定、時間可変、流量可変、一定時間可変量 (CTVV)
このマニュアルの拡張版の一定時間可変量 (CTVV) サンプリングを参照してくださ
い。
分配: 単一ボトル複合、複数ボトル複合、複数ボトル分離、サンプルあたりのボトル
数、ボトルあたりのサンプル数、サンプルあたりのボトルの組み合わせ、ボトルあた
りのサンプル数
実行モード ユーザーが入力したサンプル数で連続および非連続
保存可能なプログラム数 最大 3 つのサンプリングプログラムを保存可能
移送速度 (標準値) 0.9 m/s: 揚水高さ 4.6 m、長さ 4.9 m
3
/
8
インチビニール製取入管、21 °C、高度
1524 m の場合
液体センサ 超音波. 本体: ウルテム
®
NSF ANSI 規格 51 認定、USP Class VI 準拠
サンプルの履歴 最大 510 レコード
エアーパージ 各サンプルの前後で自動的にエアーパージ。パージ時間は取入管の長さに応じて自
動補正。
イベントログ 510
接続 電源、補助、シリアル通信、ディストリビューター、SDI-12
接液面の材質 ステンレススチール、ポリエチレン、テフロン、ウルテム、シリコン
通信 RS232ModbusSDI-12
保証 1
130
日本語
総合情報
いかなる場合も、製造元は、例えそのような損害が生じる可能性について報告を受けていたとして
も、本マニュアルに含まれるいかなる瑕疵または脱落から生じる直接的、間接的、特定、付随的また
は結果的に生じる損害に関して責を負いません。製造元は、通知または義務なしに、随時本マニュア
ルおよび製品において、その記載を変更する権利を留保します。改訂版は、製造元の Web サイト上
にあります。
安全情報
メーカーは、本製品の目的外使用または誤用に起因する直接損害、偶発的損害、結果的損害を含むあらゆる損害
に対して、適用法で認められている範囲で一切責任を負わないものとします。ユーザーは、適用に伴う危険性を
特定したり、装置が誤作動した場合にプロセスを保護するための適切な機構を設けることに関して、全責任を負
うものとします。
この機器の開梱、設定または操作を行う前に、このマニュアルをすべてよく読んでください。危険お
よび注意の注意事項に注意を払ってください。これを怠ると、オペレータが重傷を負う可能性、ある
いは機器が損傷を受ける可能性があります。
本装置に備わっている保護機能が故障していないことを確認します。本マニュアルで指定されてい
る以外の方法で本装置を使用または設置しないでください。
危険情報の使用
回避しなければ死亡または重傷につながる、潜在的または切迫した危険な状況を示します。
避けない場合、死亡事故や負傷が起こるかも知れない危険な状況を示します。
軽傷または中傷事故の原因となる可能性のある危険な状況を示しています。
回避しなければ、装置の損傷を引き起こす可能性のある状況を示します。 特に注意を要する情報。
使用上の注意ラベル
測定器上に貼付されたラベルや注意書きを全てお読みください。これを怠ると、人身傷害や装置の損
傷につながるおそれがあります。測定器に記載されたシンボルは、使用上の注意と共にマニュアルを
参照してください。
これは安全警報シンボルです。潜在的な障害を避けるためにこのシンボルのすべて安全メッセージに
従ってください。 装置上では、作業または安全情報に関しては取り扱い説明書を参照してください。
このシンボルは、指や皮膚を挟み込む可能性があることを示します。
このシンボルが付いている電気機器は、 ヨーロッパ域内または公共の廃棄処理システムで処分できま
せん。古くなったり耐用年数を経た機器は、廃棄するためにメーカーに無償返却してください。
取得認証
カナダの障害発生機器規則、IECS-003、クラス A:
日本語
131
テスト記録のサポートはメーカーにあります。
このクラス A デジタル装置はカナダの障害発生機器規則の要件をすべて満たします。
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC PART 15、クラス A 限度値
テスト記録のサポートはメーカーにあります。この機器は FCC 規則のパート 15 に準拠します。運
転は以下の条件を前提としています:
1. この装置が有害な干渉の原因とならないこと。
2. この装置が望ましくない動作の原因となる可能性のあるいかなる干渉にも対応しなければなりま
せん。
これらの規格への準拠に責任を持つ当事者による明示的承認を伴わなずにこの装置に対する改変ま
たは改造を行うと、ユーザーはこの機器を使用する権限を失う可能性があります。この装置は、FCC
規則のパート 15 に従って、クラス A のデジタル機器の制限に準拠することが試験によって確認され
ています。これらの制限は、この機器が商用の環境で使用されたときに、有害な干渉から適切に保護
することを目的に設定されています。この機器は、無線周波数エネルギーを生成および使用するもの
で、取り扱い説明書に従って取り付けおよび使用しない場合にはそれを放射する場合があり、無線通
信に対して有害な干渉を発生させる可能性があります。住宅地域における本装置の使用は有害な電
波妨害を引き起こすことがあり、その場合ユーザーは自己負担で電波妨害の問題を解決する必要があ
ります。以下の手法が干渉の問題を軽減するために使用可能です。
1. 装置から電源を取り外して、電源が干渉源かどうかを確認します。
2. 装置が干渉を受けている装置と同じコンセントに接続されている場合は、装置を別のコンセント
に接続してください。
3. 妨害を受けている装置から本装置を離します。
4. 干渉を受けるデバイスの受信アンテナの位置を変更します。
5. 上記の手法を組み合わせてみます。
製品の概要
化学的および生物学的な危険。この装置の用途が処理工程や薬液注入システムの監視であり、それらに
対して公衆衛生、公衆安全、食品/飲料の製造/加工に関する規制や監視要件が存在する場合、この装置
の使用者には、該当するすべての規制を把握して遵守する責任、および装置の異常時に関する当該規制
に従って十分かつ適切な措置を講じる責任があります。
火災の危険。本製品は、可燃性の液体を使用するように設計されていません。
SD900 Portable Sampler は液体試料を指定間隔で収集して、ボトルまたはリテイナーに保管します。
サンプラーはさまざまな水性用途、および有毒汚染物質や浮遊固体に使用します。さまざまなリテイ
ナー、ボトル、またはコンテナーを使用してサンプラーをセットアップします。 1 を参照してくだ
さい。
132
日本語
1 製品の概要
1 コンパクトベース 6 コントローラーの接続部
2 標準絶縁ベース 7 ポンプ
3 センターセクション 8 コントローラー
4 電源 9 液体センサ
5 上部カバー 10 標準絶縁ベース、21 L コンテナー用
製品コンポーネント
すべてのコンポーネントを受け取っているか確認します。 2 を参照してください。品目が不足し
ていたり損傷していたりした場合は、直ちに HACH Japan または弊社販売代理店にお問い合わせく
ださい。
2 製品コンポーネント
1 ベース (標準、コンパクト、または混合) 6 バッテリー充電器 (オプション)
2 単一のボトル使用する場合のコンポーネント (ボト
ルおよびサポートは変更可能)
7 AC 電源 (オプション)
3 コントローラーを含むセンターセクション 8 ストレーナー
4 バッテリー (オプション) 9 取入管 (ビニールまたはテフロン加工)
5 上部カバー 10 複数のボトル使用する場合のコンポーネント (
トルおよびリテイナーは変更可能)
日本語
133
設置
複合的な危険。本書のこのセクションに記載されている作業は、必ず資格のある要員が行う必要があり
ます。
閉鎖空間に関する注意事項
爆発の危険。プリエントリ試験トレーニング、換気、エントリ手順、排気/レスキュ手順および安全作業
の実行は閉じ込められたスペースに入る前に必要です。
次に示す情報は、閉鎖空間への立ち入りに関連して発生する危険およびリスクについて理解するため
のものです。
1993 4 15 日に、CFR 1910.146Permit Required Confined Spaces」に関する OSHA の最終
的な決定が法制化されました。この標準は米国の 250,000 を超える産業用地に直接影響し、閉鎖空間
で働く作業者の健康と安全を保護することを目的としています。
閉鎖空間の定義:
閉鎖空間は、次の条件が 1 つ以上該当する (またはすぐにでも該当する可能性のある) 場所または密
閉状態のことです。
酸素濃度が 19.5 % 未満また 23.5 % を超える、あるいは硫化水素 (H
2
S) 濃度が 10 ppm を超える
雰囲気
ガス、蒸気、霧、塵、または繊維による引火性のある雰囲気、または爆発の可能性がある
接触または吸引すると怪我、健康障害、または死亡を引き起こす可能性のある有毒物質が存在する
閉鎖空間は人間が滞在することを意図していません。閉鎖空間への立ち入りが制限され、既知の危険
が存在するか、危険が発生する可能性があります。閉鎖空間の例には、マンホール、煙突、パイプ、
タンク、スイッチ保管所、およびその他の同様な場所などです。
危険なガス、蒸気、霧、塵、または繊維が存在する可能性のある閉鎖的な空間または場所に立ち入る
場合は、事前に標準的な保護手順を実行する必要があります。閉鎖空間に立ち入る前に、閉鎖空間へ
の立ち入りに関連したすべての手順を検索し、参照してください。
設置
現場設置ガイドライン
爆発の危険。この装置は、危険がある場所での設置は許可されていません。
設置現場の評価については、以下のガイドラインを参照してください。
サンプラーを密閉空間に取り付ける場合は、すべての安全注意事項に従ってください。閉鎖空間に
関する注意事項 ページの 134 を参照してください。
設置場所の温度が指定範囲内であるか確認してください。仕様 ページの 129 を参照してくださ
い。
サンプラーは水平面に取り付けるか、サスペンションハーネス、サポートブラケット、またはスパ
ナーバーを使用してサンプラー吊り下げてください。サンプラーをマンホールに取り付け
ページの 135 および該当する設置説明書を参照してください。
分析遅延を小さくするために、可能な限り試料源の近くに配置してください。サンプラーの配管
ページの 137 を参照してください。
輸送速度および最大揚水高さの制限については、仕様 ページの 129 を参照してください。
134
日本語
サンプラーをマンホールに取り付け
マンホールの試料水の上にサンプラーを取り付けます。サンプラーを取り付けるには、スパナーバー
またはサポートブラケットを使用します。マンホールの内側にスパナーバーを取り付けます。スパ
ナーバーは壁に対する圧力で支えます。サポートブラケットはマンホールカバーと同じ幅です。サ
ポートブラケットはカバーの真下に取り付けて支えます。メーカーの Web サイトにある拡張版マニ
ュアルを参照してください。サンプラーの取り付け手順については、アクセサリーに付属している説
明書を参照してください。
ディストリビューターまたはフルボトルシャットオフの取り付け (オプション)
ディストリビューターまたはフルボトルシャットオフアセンブリーは通常、工場出荷時に取り付けら
れています。取り付け手順については、ディストリビューターまたはフルボトルシャットオフの説明
書を参照してください。
サンプラーの準備
サンプルボトルの洗浄
ブラシ、水、および低刺激の洗剤を使用して、サンプルボトルとキャップを洗浄します。コンテナを
真水で洗浄してから、蒸留水ですすぎます。
単一のボトルの取り付け
混合試料を 1 つ収集するには、単一のボトルを使用します。ボトルがいっぱいになると、フルボトル
シャットオフにより試料プログラムが停止します。
1. サンプルボトルを洗浄します。サンプルボトルの洗浄 ページの 135 を参照してください。
2. 3 に示しているようにサンプルボトルを取り付けます。
3 単一ボトルの取り付け
1 ポリエチレンボトル、10 L (1918) 6 ポリエチレンボトル、21 L (6494)
2 ガラスボトル、10 L (6559) 7 コンパクトベース (8975)
3 サポート (1502) 8 標準絶縁ベース (8976)
4 ポリエチレンボトル、15 L (1367) 9 標準絶縁ベース、21 L ボトル用 (8561)
5 ポリエチレンボトル、19 L (6498)
最初のボトルの取り付け (複数の試料を収集する場合)
複数の試料を個別のボトルまたは 2 本以上のボトルに収集する場合は、複数のボトルを使用します。
ディストリビューターで試料チューブを各ボトルの上に移動します。 4 に示すように、サンプラー
日本語
135
ベースにボトルを取り付けます。サンプラーベースのラベルの下に、最初の試料ボトル (番号 1) を取
り付けます。残りのボトルをラベルが示す方向に昇順に取り付けます。必要なコンポーネントの図
については、 5 ページの 137 を参照してください。
4 ボトル番号 1 の取り付け
1 ボトル番号 1 の場所 (コンパクトベース) 6 ゴム製ストラップ
2 ボトル番号 1 の場所 (ボトル数 24) 7 標準絶縁ベース、21 L ボトル用
3 ボトル番号 1 の場所 (ボトル数 8) 8 標準絶縁ベース
4 ボトル番号 2 の場所 (ボトル数 2 または 4) 9 コンパクトベース
5 リテイナー
複数のボトルの取り付け
複数のボトルを取り付ける場合は、ディストリビューターアームで試料チューブを各ボトルの上に移
動します。指定した数の試料が収集されると、試料の収集が自動的に停止します。
1. サンプルボトルを洗浄します。サンプルボトルの洗浄 ページの 135 を参照してください。
2. サンプラーベースのラベルの下に、最初の試料ボトル (番号 1) を取り付けます。最初のボトルの
取り付け (複数の試料を収集する場合) ページの 135 を参照してください。
3. 5 に示しているようにサンプルボトルを組み立てます。ボトル数が 8 本以上の場合は、最初の
ボトルをボトル 1 インジケーターの横 (時計回り方向) に配置します。
4. ボトルアセンブリーをサンプラーに配置します。ボトル数が 8 本以上の場合は、下部トレイの溝
にワイヤを揃えます。
136
日本語
5 複数のボトルの取り付け
1 ポジショナー/リテイナー (2347) 8 ガラスボトル、3.8 L (4x) (2216) またはポリエチレ
ンボトル、3.8 L (4x) (2217)
2 リテイナー (1422) 9 ガラスボトル、1.9 L (8x) (1118) またはポリエチレ
ンボトル、2.3 L (8x) (657)
3 リテイナー (2190) 10 ポリエチレンボトル、1 L (24x) (737)
4 リテイナー (2189) 11 ガラスボトル、350 mL (24x) (732)
5 ガラスボトル、950 mL (8x) (2348) 12 コンパクトベース (8975)
6 ポリエチレンボトル、575 mL (24x) (1369) 13 標準ベース (8976)
7 ガラスボトル、3.8 L (2x) (2214) またはポリエチレ
ンボトル、3.8 L (2x) (2215)
サンプラーの配管
ストレーナーを (表面付近や下部ではなく) 試料水流の中央に取り付けて、代表的な試料が収集される
ようにします。 6 を参照してください。取入管の取り付け手順については、 7 を参照してくださ
い。
1. 7 に示しているようにチューブをサンプラーに接続します。
:
テフロンライニングチューブを使用する場合は接続キット
2186
を使用します。
2. 取入管とストレーナーを試料源の本流に設置し、水が攪拌されて十分に混合されるようにします。
取入管はできるだけ短くしてください。
サンプルを採取するたびにチューブが完全に排水されるよう、取入管の傾斜が保たれるように
してください。
:
傾斜を保つことができない場合、またはチューブが加圧される場合は、液体センサを無効にしてくださ
い。試料量を手動で調整します。
サンプル量の手動校正
ページの
147
を参照してください。
取入管がつぶれていないことを確認してください。
日本語
137
6 器具の取り付け
1 ストレーナー 3 取入管
2 揚水高さ
7 取入管の取り付け
配線
コントローラーの接続部
電気ショックの危険。外部接続された機器には、該当する国の安全標準評価が必要です。
8 に、コントローラーの接続部を示します。
138
日本語
8 コントローラーのコネクタ
1 補助デバイス 4 SDI-12 デバイスオプション
2 電源 5 ディストリビューター/フルボトルシャットオフ
3 シリアル通信
流量計の接続
サンプルの流量が指定値を上回ったか下回った場合にサンプラーを開始または停止させるには、コン
トローラーに流量計を接続します。
用意するもの:
Sigma 流量計用の多目的フルケーブル (またはモデル 980 流量計用の 980 ハーフケーブル)
追加の接続用のオプションのスプリッター。複数のスプリッターを直列接続できます。
1. ケーブルの一方の端を流量計に接続します。モデル 980 流量計の場合は、モデル 980 流量計の取
扱説明書を参照してください。
2. ケーブルのもう一方の端をコントローラーの補助デバイスコネクタに接続します。
:
流量計のケーブルが
6
ピンの場合は、
6
ピン
-> 7
ピン変換アダプターケーブルを使用してください。
PC または通信ネットワークの接続
データを転送したりサンプラーを設定したりするには、PC または通信ネットワークをコントローラ
ーに接続します。
用意するもの:
シリアルケーブル、7 ピン RS232 -> DB-9 ( 9 を参照)。接続: B -> 5 (信号接地)D -> 3 (RCD)
F -> 2 (TXD)G (接地)
1. シリアルケーブルの一方の端を通信デバイスまたはネットワークに接続します。
2. ケーブルのもう一方の端をコントローラーのシリアル通信コネクタに接続します。
9 シリアル通信ケーブル
日本語
139
サンプラーを電源に接続
感電死の危険。この装置を屋外または湿っている可能性がある場所で使用する場合は、主電源との接続
に漏電回路安全装置 (GFCI/GFI)を使用する必要があります。
火災の危険。電力線に 15 A の回路ブレーカーを取り付けます。回路ブレーカーは、機器の近くにあれ
ば、ローカルの電源切断装置になります。
感電死の危険。保護アース接地(PE)接続が必要です。
感電死の危険。ローカル電源切断へのアクセスが容易であることを確認します。
電気ショックの危険。電源はポンプサイクル間隔が短すぎる場合に過熱することがあります。
ポンプサイクル時間によって電源が過熱状態にならないことを確認します。ポンプのサイクル時間の計算
ページの 141 を参照してください。
サンプラーをバッテリーまたは AC 電源に接続します。 下図に示した手順を参照してください。
140
日本語
ポンプのサイクル時間の計算
電源が過熱しないように、ポンプはサンプルサイクル間で一定時間以上オフになっている必要があり
ます。
1. サンプルサイクル中のポンプの合計連続運転時間を計算します。計算には、プレパージ、取入管
洗浄、サンプル、サンプル採取リトライ、ポストパージというすべての段階を含めます。
2. 10 を参照して、サンプルサイクル間でポンプがオフになっていなければならない最小時間を見
つけます。
3. サンプルサイクル間でポンプがこの最小時間にわたってオフになるように、サンプラーのプログ
ラムのペースまたは時間間隔が設定されていることを確認します。
10 ポンプのオン/オフ時間 (50 °C )
1 ポンプがオフになっていなければならない時間 (
単位)
3 : ポンプが 5 分間連続的にオンになる場合、ポン
プは 15 分間オフになる必要があります。
2 ポンプが連続的にオンになっている時間 (分単位)
電源の投入
電源ボタンを押すと、電源のオン/オフを切り替えることができます。電源 (AC 電源またはバッテリ
ー電源) が正しく取り付けられていることを確認してください。
ユーザーインターフェースとナビゲーション
コントローラーのキーパッドを 11 に示します。[MENU (メニュー)] キーは、サンプリングプログ
ラムの設定、コントローラーの設定、診断テストの実行に使用します。矢印キー、[ENTER (入力)]
ー、および [BACK (戻る)] キーは、メニューのスクロール、項目の選択、値の入力に使用します。デ
ィスプレイ画面の下または横に矢印が表示されている場合は、選択可能なオプションがさらにあるこ
とを示しています。
日本語
141
11 コントローラーのキーパッド
1 POWER (電源) 5 MENU (メニュー) 9 STATUS (状態)
2 VOLUME CALIBRATION (量の
校正)
6 ENTER (入力) 10 RUN/HALT PROGRAM (プロ
グラムの実行/中止)
3 STOP (停止) 7 BACK (戻る) 11 MANUAL OPERATION (手動操
)
4 LED 8 矢印キー
操作
システム設定の指定
1. [MENU (メニュー)] を押します。
2. [SYSTEM SETUP (システム設定)] に進みます。
3. 該当するオプションを変更します。
オプション 説明
Time/date setup (日時設定) 時刻 (24 時間形式) と日付を設定します。
Communication (通信) シリアルポートのボーレート (192003840057600または 115200) とプロ
トコル (Modbus RTU または ASCII) を選択します。
Setup base (ベースの設定) サンプラーベースを選択します (全天候型冷蔵式、冷蔵式、携帯型標準、また
は携帯型コンパクト)
Language(言語) コントローラーの言語を選択します。
Liq sensor cal (液体センサ
校正)
液体センサを校正するか、デフォルトに設定します。
Set contrast (コントラスト
設定)
LCD 画面のコントラストを調整します。
Pump duty cycle (ポンプの
デューティサイクル)
ポンプのデューティサイクルを変更します。範囲: 50 100 % (デフォルトは
100 %)
142
日本語
オプション 説明
Password setup (パスワー
ド設定)
プログラム設定メニューおよびシステム設定メニューへのアクセスを制御す
るためのパスワードを設定します。[ENABLE (有効)] > [YES (はい)] を選択し
て新しいパスワードを設定するか、[ENABLE (有効)] > [NO (いいえ)] を選択し
て現在のパスワードを有効にします。パスワードを初めて設定する場合は、
在のパスワードに 900900 を使用してください。
Tubing life (チューブ寿命) ポンプチューブの変更が必要になったときにそれを知らせる通知機能を設定
します。ポンプのサイクルカウントを 0 にリセットするには、[ENABLE (
)] > [RESET CYCLES (サイクルのリセット)] に進みます。通知機能に使用
するポンプのサイクル数を変更するには、[ENABLE (有効)] > [CYCLE LIMIT
(サイクル上限)] に進みます。ポンプのサイクルカウントの現在の値を表示す
るには、[ENABLE (有効)] > [STATUS (ステータス)] に進みます。
SDI–12 (オプション) SDI–12 デバイスの動作を設定します。このオプションはデバイスが検出され
た場合のみ表示されます。
サンプリングプログラムの設定
サンプリングプログラムには、サンプルを自動的に採取するのに必要なすべてのパラメーターが含ま
れています。サンプルは一定の間隔で、またはサンプリングプログラムが完了したときに採取されま
す。
1. [MENU (メニュー)] を押します。
2. [PROGRAM SETUP (プログラム設定)] > [MODIFY ALL (すべて修正)] に進みます。最初の画面が
表示されます。
3. 各パラメーターの値を選択または入力します。左または右へ移動するには、左または右矢印キー
を使用します。値を変更するには上または下矢印キーを使用します。矢印キーを使用して、必要
に応じて単位 (gal または mL) を変更します。
オプション 説明
Bottles (ボトル) ボトル数量: サンプラーに装填されているボトルの数 (1248、または 24)
Full BTL shutoff (フルボトルシャットオフ): Enabled (有効)/Disabled (無効)。有効に
した場合、コントローラーはボトルのオーバーフロー状態を確認します。
Bottle volume (ボトル容量): 各ボトルの容量。範囲: 50 65,000 mL
Intake tubing (取入管) Tubing length (チューブの長さ): ストレーナーから液体センサまでの取入管の長さ。
範囲: 100 3000 cm。正確なサンプル量を得るためには正確な長さを入力する必
要があります。
Intake tube type (取入管のタイプ): 取入管の径と材質 (1/4 インチビニール、3/8 イン
チビニール、または 3/8 インチテフロン)
Program delay (プロ
グラム遅延)
Enabled (有効)/Disabled (無効): 有効にすると、サンプリングプログラムは指定され
た時間またはカウント数に達したときに開始されます。Date and time (日時): プロ
グラムが開始する日時 (24 時間形式)Counts (カウント): プログラムが開始する、
フローメーターからのカウント数またはパルス数 (1 9999 カウント)。サンプル
のペースを時間に設定した場合、プログラム遅延は無効になります。
Sample
pacing/collection (
ンプルのペース/採取)
時間ベース - 時間ベースの採取 ページの 144 を参照してください。
: CTVV
を使用する場合
:
液体センサは必ず有効にします。
サンプルの分配は複合です。
実行モードは非連続です。
最初のサンプルは時間間隔の後に採取されます。
フローソースは常に
4
20 mA
入力です。
流量ベース - 流量ベースの採取 ページの 145 を参照してください。
日本語
143
オプション 説明
Sample distribution
(サンプルの分配)
Deliver samples to all (すべてにサンプルを供給) > YES (はい)
各サンプルがすべてのボトルに供給されます。End after last (最後のサンプルの後
に終了): プログラムは最後のサンプルの後に終了します (1 999)Continuous (
): プログラムは手動で停止するまで続行されます。
Deliver samples to all (すべてにサンプルを供給) > NO (いいえ)
試料は一部のボトルに供給されます Samples/bottle (サンプル/ボトル): 各ボトル
に採取するサンプルの数 (1 999)Bottles/sample (ボトル/サンプル): 同じサンプ
ルが含まれるボトルの数。
Liquid sensor (液体セ
ンサ)
有効または無効。無効の場合、サンプル量は時間で校正する必要があります。
Sample volume (サン
プル量)
各サンプルの量 (mL 単位)。ボトル/サンプルモードを使用した場合、指定した量の
サンプルが各ボトルに入れられます (10 10,000 mL)。サンプルの量がボトルの容
量よりも多くならないようにしてください。サンプルの量は 10 mL 単位で丸められ
ます。
Intake rinses (取入管
洗浄)
サンプルが採取される前に取入管が洗浄される回数です (0 3)
Sample retries (サン
プル採取リトライ)
サンプリングが失敗した後に、サンプリングを再試行する回数です (0 3)
Site ID (サイト ID) サンプリング場所の名前です (最大 12 文字)。プログラムを保存した場合、サイト
ID はプリセット名として使用されます。
Advanced sampling
(高度なサンプリング)
高度なサンプリングメニューに進みます。必要に応じて、[Advanced Sampling (高度
なサンプリング)] メニューのオプションを設定します。
Run program (プログ
ラム実行)
基本的なサンプリングプログラムの作成を終了します。プログラムの開始またはキ
ャンセルを求めるメッセージがサンプラーに表示されます。
時間ベースの採取
1. 各オプションの値を選択または入力します。
オプション 説明
Pacing interval (ペース
間隔)
サンプルを一定の時間間隔で採取します。時間と分 (0:01 999:00) を指定しま
す。
Take first sample (最初
のサンプルを採取)
プログラムは直ちに開始されるか、または最初の時間間隔後に開始されます。
Variable volume (可変
採水量)
サンプル量を変化させるかどうかを指定します (流量ベース)。変化させる場合は、
可変量を使用する ページの 144 を参照してください。変化させない場合は、可変
量を使用しない ページの 145 を参照してください。
可変量を使用する
サンプルペースは一定時間可変量 (CTVV) です。
1. 各オプションの値を選択または入力します。
オプション 説明
単位選択 使用する流量単位を選択します。
Map 4 mA input (4 mA 入力マッ
ピング)
4 mA 入力に相当する流量を入力します。
Map 20 mA input (20 mA 入力マ
ッピング)
20 mA 入力に相当する流量を入力します。
144
日本語
オプション 説明
Average flow rate (平均流量) サイトの平均流量を指定します (1 999,999)。単位は前述の手順で選
択した単位です。
Time volume desired (必要な時
間量)
サンプルプログラム中に採取される合計量を指定します (10
10,000 ml)
Collection period (採取期間) 合計採取期間を時間および分単位で指定します (0:01 999:00)
最小サンプル量 この最小量を採取できる流量がない場合、サンプルはスキップされます
(10 10,000 ml)
可変量を使用しない
1. オプションの値を選択または入力します。
オプション 説明
最初のサンプルを採取します。 プログラムが直ちに開始されるか、または最初の時間間隔後に開始されるか
を選択します。
流量ベースの採取
流量ベースのサンプリングのフローソースを指定します。
カウント - フローソースのカウントを使用する ページの 145 を参照してください。
4 20 mA 入力 - フローソースの 4 20 mA 入力を使用します。 ページの 145 を参照してくだ
さい。
フローソースのカウントを使用する
1. 各オプションの値を選択または入力します。
オプション 説明
Take sample every (サンプル採
取間隔)
一定の流量間隔でサンプルを採取します。カウント (1 9999) を指定し
ます。
Overrride time (サンプル採取間
)
流量が非常に少ない場合にサンプルを採取します。
Time (時間) サンプル間の最大時間 (0:01 999:00)。各サンプルが採取された後に、
タイマーが再度開始されます。
Take first sample (最初のサンプ
ルを採取)
プログラムは直ちに開始されるか、または最初の時間間隔後に開始されま
す。
フローソースの 4 20 mA 入力を使用します。
1. 各オプションの値を選択または入力します。
オプション 説明
単位選択 使用する流量単位を選択します。
Map 4 mA input (4 mA
入力マッピング)
4 mA 入力に相当する流量を入力します。
Map 20 mA input
(20 mA 入力マッピング)
20 mA 入力に相当する流量を入力します。
Take sample every (サン
プル採取間隔)
サンプルを採取した後の合計流量を入力します。
割り込み時間 選択すると、流量が極端に低い場合に採取するサンプルの押し出しが有効になり
ます。
日本語
145
オプション 説明
Time (時間) サンプル間の最大時間 (0:01 999:00)。各サンプルが採取された後に、タイマー
が再度開始されます。
サンプル量 サンプルごとに採取する量を入力します。ボトル/サンプルモードを選択した場
合、指定した量のサンプルが各ボトルに入れられます (100 10,000 mL)。可変
量を選択した場合、このオプションはスキップされます。
デフォルト設定の復元
復元オプションを実行すると、すべてのプログラムパラメーターがデフォルト設定に設定されます。
1. [MENU (メニュー)] を押します。
2. [PROGRAM SETUP (プログラム設定)] > [RESTORE (復元)] に進みます。
3. [YES (はい)] を選択します。
プログラムの起動/停止
[RUN/HALT PROGRAM (プログラムの実行/中止)] キーを使用して、プログラムを起動または停止し
ます。最大 12 個のユーザー定義の起動/停止日時を設定できます。プログラムまたはシステムの設
定を変更するには、プログラムは停止する必要があります。
1. [RUN/HALT PROGRAM (プログラムの実行/中止)] を押します。
2. いずれかのオプションを選択します。
オプション 説明
開始 現在ロードされているプログラムが開始します。ステータスが [RUNNING (実行中)]
に変わります。
Halt (中止) プログラムが一時的に停止します。ステータスが [HALTED (中止)] に変わります。
Resume (再開) 中止された位置からプログラムが再開されます。ステータスが [RUNNING (実行中)]
に変わります。
Start over (やり直し) プログラムが最初から開始されます。ステータスが [RUNNING (実行中)] に変わり
ます。
End program (プログ
ラム終了)
現在実行中のプログラムが停止します。ステータスが [COMPLETE (完了)] に変わり
ます。
手動操作
グラブサンプルを採取したり、ディストリビューターアームを移動したり、ポンプを作動させたりす
るには、手動操作を使用します。追加情報はこのマニュアルの拡張バージョンを参照してください。
量の校正
サンプル量は、液体センサを使用して自動的に校正することも、手動で校正することもできます。液
体センサを使用して量の校正を行った場合、プログラムされているすべての量を正確に測定するよう
にセンサが調整されます。液体センサが無効になっている場合は、サンプル、洗浄、および雨水の量
を手動で校正する必要があります。
標準的な用途の場合は、量の校正に水道水を使用することができます。サンプルの組成が標準的な水
サンプルと大幅に異なる場合は、採取する液体でサンプラーを校正します。校正を実行する前に、
べてのプログラムを停止する必要があります。
液体センサの校正
液体センサの精度は、サンプリングされた液体の種類によって異なります。たとえば、濁度の高いサ
ンプルでは、清浄な水の場合よりも量の精度が低くなることがあります。サンプル水で液体センサを
校正して、量の精度を向上させることができます。
146
日本語
1. 取入管をサンプル水に入れます。
2. [MENU (メニュー)] を押します。
3. [PROGRAM SETUP (プログラムの設定)] > [MODIFY SELECTED (選択修正)] > [LIQUID
SENSOR (液体センサ)] > [ENABLE (有効)] に進み、液体センサが有効になっていることを確認し
ます。
4. [SYSTEM SETUP (システム設定)] > [LIQ SENSOR CAL (液体センサ校正)] に進みます。
5. [PERFORM CAL (校正実行)] を選択します。
6. [START (開始)] を選択します。ポンプが逆転して、取入管がパージされます。その後、ポンプは
正転します。
7. 取入管のサンプルの流れを見て、サンプルが液体センサ内を流れていることを確認します。
8. サンプルが液体センサ内を流れ、気泡が存在しなくなったら、[STOP (停止)] を押します。取入管
が自動的にパージされます。これで液体センサの校正は完了です。
9. [DONE (完了)] を選択して終了するか、[REPEAT (繰り返し)] を選択して再度校正を実行します。
デフォルト校正の復元
液体センサの量の校正は、デフォルトの工場出荷設定に戻すことができます。
1. [PROGRAM SETUP (プログラム設定)] > [MODIFY SELECTED (選択修正)] > [LIQUID SENSOR
(液体センサ)] > [ENABLE (有効)] を選択して、液体センサが有効になっていることを確認します。
2. [VOLUME CALIBRATION (量の校正)] を押して、[RESET CAL (校正リセット)] を選択します。量
の校正がデフォルト設定に設定されます。value has been reset to 0 (値が 0 にリセットされまし
)」というメッセージが表示されます。
液体センサを使用したサンプル量の校正
サンプル量を微調整するには、液体センサが有効な状態でサンプル量を校正します。
1. [PROGRAM SETUP (プログラムの設定)] > [MODIFY SELECTED (選択修正)] > [LIQUID
SENSOR (液体センサ)] > [ENABLE (有効)] に進み、液体センサが有効になっていることを確認し
ます。
2. 取入管をサンプル水または水道水に入れます。
:
最高の精度を得るためにもサンプル水を使用してください。
3. ポンプの出口側から出ているチューブをメスシリンダーに入れます。
4. [VOLUME CALIBRATION (量の校正)] を押して、[CALIBRATION (校正)] を選択します。
5. [START (開始)] を選択します。ポンプが逆転して、取入管がパージされます。その後、ポンプが
正転して、サンプリングプログラムに指定されているサンプル量が採取されます。ポンプが逆転
して、取入管がパージされます。
6. サンプルの採取が完了したら [DONE (完了)] を選択します。メスシリンダー内のサンプル量と、
プログラム設定メニューに入力したサンプル量を比較します。
7. 採取された量が現在のプログラムに指定されているサンプル量と異なる場合は、実際に採取され
た量を入力します。[ENTER (入力)] を押します。プログラムされているすべての量が正確に測
定されるようにセンサが調整されます。
8. [START (開始)] を選択してサンプリングプログラムを開始するか、[CANCEL (キャンセル)] を選
択して量の校正メニューを終了します。
サンプル量の手動校正
液体センサが無効な場合は、すべての量を時間によって手動校正する必要があります。サンプル量
は、現在のプログラムに指定されている量で校正されます。プログラムでサンプル量を変更した場合
は、新しいサンプル量で再度校正する必要があります。
1. [PROGRAM SETUP (プログラムの設定)] > [MODIFY SELECTED (選択修正)] > [LIQUID
SENSOR (液体センサ)] > [DISABLE (無効)] を選択して、液体センサが無効になっていることを確
認します。
2. 取入管をサンプル水または水道水に入れます。
日本語
147
3. ポンプの出口側から出ているチューブをメスシリンダーに入れます。
4. [VOLUME CALIBRATION (量の校正)] を押します。複数の量が表示された場合は、校正するいず
れかの量を選択します。
5. [START (開始)] を選択します。ポンプが逆転して、取入管がパージされます。その後、ポンプが
正転して、サンプルの採取が開始されます。
6. 選択した量または洗浄位置でポンプを停止します。
サンプル量: メインプログラムに指定されている量が採取されたら [STOP (停止)] を押します。
洗浄: 取入管内の液体が液体検出器に達したら [STOP (停止)] を押します。
雨水量: 雨水プログラムに指定されている量が採取されたら [STOP (停止)] を押します。
7. [DONE (完了)] を選択して次の量の校正に進むか、[REPEAT (繰り返し)] を選択して再度校正を行
います。
8. すべての量の校正が完了したら、[DONE (完了)] を選択します。
9. 出口側のチューブをサンプラーのチューブ継手に接続します。
10. [START (開始)] を選択して、サンプリングプログラムを開始するか、[CANCEL (キャンセル)]
選択して終了します。
サンプル量の確認
サンプル量が正確であるかどうかを確認するには、グラブサンプルを採取します。量を確認する目的
では校正に戻らないでください。そうすると、校正開始時に量の補正値がゼロにリセットされるから
です。
1. [MANUAL OPERATION (手動操作)] を押します。
2. [GRAB SAMPLE (グラブサンプル)] を選択します。
3. 取入管をサンプル水に入れます。
4. ポンプの出口側から出ているチューブをメスシリンダーに入れます。
5. 液体センサが有効になっている場合は、確認する量を入力します。
6. 液体センサが無効になっている場合は、プログラムに指定した量を入力します。
7. [ENTER (入力)] を押します。ポンプのサイクルが開始されます (パージ、サンプリング、パー
)
8. メスシリンダー内の採取された量と想定していた量を比較します。採取された量が正しくない場
合は、量の校正を再度実行します。
メンテナンス
複合的な危険。本書のこのセクションに記載されている作業は、必ず資格のある要員が行う必要があり
ます。
感電死の危険。メンテナンスまたはサービス活動を行う前に装置から電源を切り離します。
生物学的危険への曝露。試料ボトルおよびサンプラーコンポーネントと接触している間は安全処理手
順に従ってください。
148
日本語
複合的な危険。メンテナンス作業後には、機器が安全かつ正しく作動することを技術者が確認する必要
があります。
装置の清掃
火災の危険。装置の洗浄に、可燃性の洗浄剤を使用しないでください。
コントローラーコンパートメントヒーターは液体類を使って清掃しないでください。
水ではコントローラーとポンプを十分に清掃できない場合は、コントローラーの接続を外して、コン
トローラーをサンプラーから取り外してください。コントローラーとポンプを完全に乾かしてから、
部品を取り付けて作動させてください。
サンプラーは次の手順で清掃します:
サンプラー筐体湿った布と中性洗剤で内部および外部の表面を清掃してください。研磨剤入り
洗浄剤や溶剤は使用しないでください。
サンプラーボトルブラシと水および中性洗剤でボトルを清掃してください。真水ですすいでか
ら、蒸留水ですすいでください。 必要に応じて、加圧滅菌器でガラスボトルを清掃してください。
乾燥剤の交換
コントローラー内の乾燥剤カートリッジは、湿気を吸収し、腐食を防止します。点検窓を通して乾燥
剤の色を確認できます ( 12)。新品の乾燥剤の色は黄色です。緑色になったら乾燥剤を交換してく
ださい。
1. 乾燥剤カートリッジを回して取り外します ( 12)
2. プラグを取り外して、使用済みの乾燥剤を廃棄します。
3. 乾燥剤カートリッジに新品の乾燥剤を入れます。
4. プラグを取り付けます。
5. O リングにグリスを塗布します。
6. 乾燥剤カートリッジをコントローラーに取り付けます。
12 乾燥剤カートリッジ
1 プラグ 3 O リング
2 乾燥剤カートリッジ 4 乾燥剤点検窓
日本語
149
ポンプのメンテナンス
作業に際して、機器に指や皮膚を挟み込む可能性があります。メンテナンスまたは修理作業を完了する
前に、装置の電源を切ってください。
ポンプチューブの交換
メーカーが供給したチューブ以外のチューブを使用すると、機械部品が過度に摩耗したり、ポンプの性能が低下
したりすることがあります。
ポンプチューブのローラーが当たる部分の摩耗を点検してください。通知メッセージが表示された
場合、またはチューブに摩耗の形跡が認められた場合は、チューブを交換してください
必要条件:
ポンプチューブ - プレカットまたはバルク 4.6 m または 15.2 m
1. コントローラーの電源を切り離します。
2. バルクチューブを使用する場合は、チューブをカットして、位置決め用のドットを付けます。
ーカーの Web サイトにある拡張版マニュアルを参照してください。
3. 図の番号に従ってポンプチューブを取り外します。
4. ポンプハウジングの内部とローラーに付着しているシリコン残留物を取り除きます。
5. 図の番号に従って新品のポンプチューブを取り付けます。
150
日本語
ローターの清掃
低刺激の洗剤を使用して、ローター、ポンプチューブ溝、およびポンプハウジングを清掃します。
ンプチューブの交換 ページの 150 とそれに続いて図示されている手順に従って取り付けてくださ
い。
日本語
151
ディストリビューターアームチューブの交換
ディストリビューターアームは、複数ボトルのサンプリング時に各ボトルの上を移動します。ディス
トリビューターアームのチューブが摩耗したら交換してください。ディストリビューターとディス
トリビューターアームに適したチューブを使用してください。
:
ディストリビューターチューブはポンプチューブと異なります。ディストリビューターアセンブリにポンプチ
ューブを取り付けると、ディストリビューターが損傷することがあります。また、ディストリビューターアームは
容易に動かせないので、試料が失われることもあります。
1. ディストリビューターアームと、センターセクションの天井からチューブを取り外します。
2. ディストリビューターアームに新品のチューブを取り付けます。チューブが 13 に示された寸
法に合っているか確認します。
3. チューブのもう一方の端をセンターセクションの天井にある継手に差し込みます。
4. ディストリビューターの診断テストを行って、正しく動作することを確認します。
13 ディストリビューターアセンブリ
1 ノズル 3 ディストリビューターモーター 5 ディストリビューターチューブ
2 ディストリビューターアーム 4 シャフト
ヒューズの交換
ヒューズはユーザーが交換可能な部品ではありません。技術サポートにお問い合わせください。ヒ
ューズが切れた場合は、サンプラーに問題が発生していて、点検整備が必要な可能性があります。
152
日本語
목차
사양 페이지의 153 작동 페이지의 166
일반 정보 페이지의 155 유지관리 페이지의 172
설치 페이지의 158
추가 정보
제조 회사의 사이트에서 추가 정보를 확인할 있습니다.
사양
사양은 사전 예고 없이 변경될 있습니다.
SD900 휴대용 샘플 채취기
사양 표준형 베이스 소형 베이스 복합형 베이스
규격 지름: 50.5 cm(19.9 in.)
높이: 69.4 cm(27.3 in.)
지름: 44.1 cm(17.4 in.)
높이: 61 cm(24 in.)
지름: 50.28 cm(19.8 in.)
높이: 79.75 cm(31.4 in.)
무게 15 kg(35.6 lb), 1 L 폴리에틸렌 24
포함
14.8 kg(32.6 lb), 10 L(2.5 gal) 폴리에
틸렌 용기 1 포함
12.2 kg(27 lb),
575 mL(19.44 oz) 폴리에틸렌
24 포함
12.9 kg(28.3 lb), 10 L(2.5 gal)
리에틸렌 용기 1 포함
15 kg(36 lb),
950 mL(32.12 oz) 폴리에틸렌
12 포함
케이
3 구조의 충격 방지 ABS, 2.54 cm(1 in.) 이중 벽구조 베이스 단열 처리 - 얼음에 직접 접촉 가능한
샘플
온도
0~60 °C(32~140 °F)
여과
표준 규격의 316 스테인리스 스틸, 저심도 채취용 고속 또는 저속 프로필, Teflon
®
또는 표준 규격의 316
테인리스 스틸
샘플
흡입
배관
9.5 mm(3/8 in.) I.D 비닐 또는 Teflon 라이닝 폴리에틸렌
샘플
1 L(0.26 gal) 폴리에틸렌 /또는
350 mL(11.83 oz) 유리병 24
2.3 L(0.6 gal) 폴리에틸렌 /또는
1.9 L(0.5 gal) 유리병 8
3.8 L(1 gal) 폴리에틸렌 /또는
3.8 L(1 gal) 유리병 4
3.8 L(1 gal) 폴리에틸렌 /또는
3.8 L(1 gal) 유리병 2
21 L(5.5 gal) 또는 15 L(4 gal) 폴리에틸
합성 용기 또는 20 L(5.25 gal) 폴리
에틸렌 또는 10 L(2.5 gal) 폴리에틸
또는 10 L(2.5 gal) 유리병 1
575 mL(19.44 oz) 폴리에틸렌
24
950 mL(32.12 oz) 유리병 8
10 L(2.5 gal) 폴리에틸렌 1
10 L(2.5 gal) 유리병 1
21 L(5.5 gal) 폴리에틸렌 1
SD 컨트롤러
사양 세부 사항
치수 (W x H x D)
29.2 x 17.1 x 26.4 cm (11½ x 6¾ x 10
3
/
8
in.)
무게
4.2 kg (9.26 lb)
한글
153
사양 세부 사항
케이스 PC/ABS 혼합, NEMA 4X, 6, IP 67
전원 요구 사항 8754500 전원공급장치로 공급되는 15 VDC, 내부 전원공급장치로 공급되는 15 VDC
과부하 보호 펌프용 7 A, DC 라인 퓨즈
작동 온도
0-50°C(32-122°F)
보관 온도
–30-60 °C (–22-140 °F)
보관/작동 습도 100% 응결
펌프 튜브 연동 고속, 스프링 장착식 Nylatron 롤러
펌프 케이스
IP37
펌프 배관 9.5 mm ID x 15.9 OD mm (
3
/
8
in. ID x
5
/
8
in. OD) 실리콘
펌프 배관 수명 다음 조건에서 20,000 샘플 사이클: 1 L (0.3 gal) 샘플 볼륨, 1 린스, 6 페이싱 주기,
4.9 m (16 ft) 길이
3
/
8
in. 흡입 튜브, 4.6 m (15 ft) 수직 양정, 21 °C (70 °F) 샘플 온도
샘플 수직 양정 다음 조건에서 최대 8.5 m (28 ft): 8.8 m (29 ft) 길이
3
/
8
-in. 비닐 흡입 튜브, 해발,
20–25 °C (68–77 °F)
펌프 유량 4.8 L/min (1.25 gpm),
3
/
8
-in. 흡입 튜브를 사용하고 1 m (3 ft) 수직 양정 (일반)
시료 볼륨 10-10,000 mL (3.38 oz-2.6 gal) 범위에서 10 mL (0.34 oz) 단위로 프로그래밍 가능
시료 볼륨 반복성 (
)
다음 조건에서 무교정 액체 검출 200 mL 샘플 볼륨의 ±5%: 4.6 m (15 ft) 수직 양정,
4.9 m (16 ft) 길이
3
/
8
-in. 비닐 흡입 튜브, 단일 용기, 용기 경우 중단, 실온 1524 m
(5000 ft) 고도
시료 볼륨 정확도 (
)
다음 조건에서 무교정 액체 검출 200 mL 샘플 볼륨의 ±10%: 4.6 m (15 ft) 수직 양정,
4.9 m (16 ft) 길이
3
/
8
-in. 비닐 흡입 튜브, 단일 용기, 용기 경우 중단, 실온 1524 m
(5000 ft) 고도
샘플링 모드 속도: 시간 고정, 유량 고정, 시간 가변, 유량 가변, CTVV(일정 시간 동안 변화 가능 볼륨)
CTVV(일정 시간 동안 변화 가능 볼륨) 설명서의 확장 버전을 참조하십시오.
분산: 단일 병에 합성, 여러 병에 합성, 여러 병에 분리, 샘플 , 샘플, 샘플
조합, 샘플
실행 모드 연속식 또는 비연속식 (사용자가 샘플 입력)
다중 프로그램 최대 3 개의 샘플 채취 프로그램 저장
이송 속도 (일반) 다음 조건에서 0.9 m/s (2.9 ft/s): 4.6 m (15 ft) 수직 양정, 4.9 m (16 ft) 길이
3
/
8
-in. 비닐
튜브, 21 °C (70 °F) 1524 m (5000 ft) 고도
액정 센서 초음파. 본체: Ultem
®
NSF ANSI 표준 51 승인, USP Class VI 준수
샘플 기록 최대 510 기록
공기 정화 샘플 전후에 자동으로 공기 정화. 지속 시간이 자동으로 흡입 튜브 길이 변화를 보정.
이벤트 로그 510 기록
연결 전원, 보조, 시리얼 통신, 디스트리뷰터, SDI-12
습식 재질 스테인리스 스틸, 폴리에틸렌, Teflon, Ultem, 실리콘
통신
RS232, Modbus, SDI-12
보증 1
154
한글
일반 정보
제조업체는 설명서에 존재하는 오류나 누락에 의해 발생하는 직접, 간접, 특수, 우발적 또는 결과적
손해에 대해 어떠한 경우에도 책임을 지지 않습니다. 제조업체는 설명서와 여기에 설명된 제품을
언제라도 통지나 추가적 책임 없이 변경할 있습니다. 개정본은 제조업체 사이트에서 확인할
있습니다.
안전 정보
제조사는 제품의 잘못된 적용 또는 잘못된 사용으로 인한 직접, 우발적 또는 간접적 손해에 국한하지 않는 모든
손해에 대한 어떠한 책임도 지지않으며, 관계 법령이 최대한 허용하는 손해에 관한 면책이 있습니다. 사용자는
용상 중대한 위험을 인지하고 장비 오작동이 발생할 경우에 대비하여 적절한 보호 장치를 설치하여야 합니다.
장치 포장을 풀거나 설치하거나 작동하기 전에 설명서를 모두 읽으십시오. 모든 위험 주의사항
설명에 유의하시기 바랍니다. 이를 지키지 않으면 사용자가 중상을 입거나 장치가 손상될 있습니
.
장치의 보호 기능이 손상되지 않도록 설명서에서 설명하는 방법이 아닌 다른 방법으로 장치
사용하거나 설치하지 마십시오.
위험 정보 표시
방지하지 않을 경우 사망 또는 심각한 부상이 일어나는 잠재적 또는 즉각적 위험 상황을 의미합니다.
피하지 않을 경우에 사망이나 심각한 부상을 유발할 있는 잠재적 위험이나 긴급한 위험 상황을 나타냅니다.
경미하거나 심하지 않은 부상을 초래할 있는 잠재적으로 위험한 상황을 경고합니다.
피하지 않으면 기기에 손상을 일으킬 있는 상황을 나타냅니다. 특별히 강조할 필요가 있는 정보.
주의 경고
기기에 부착된 모든 라벨 태그를 참조하시기 바랍니다. 지침을 따르지 않을 경우 부상 또는 기기
손상이 발생할 있습니다. 기기에 있는 기호는 주의사항에 대한 설명과 함께 설명서에서 참조합니
.
이는 안전 경고 심볼입니다. 잠재적인 부상 위험을 방지할 있도록 기호를 따라 모든 안전 메시지
준수하십시오. 기기에 안전 기호가 부착되어 있는 경우 작동 안전 정보에 대해서는 작동 설명서를
참조하십시오.
심볼은 잠재적인 압착 위험을 나타냅니다.
심볼이 표시된 전기 장비는 유럽 공공 폐기 시스템에 따라 폐기할 없습니다.
인증
캐나다 무선 간섭 유발 장치 규정, IECS-003, 등급 A:
보조 테스트 기록은 제조업체가 제공합니다.
한글
155
등급 A 디지털 장치는 캐나다 간섭 유발 장치 규제의 모든 요구조건을 만족합니다.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC Part 15, Class "A" 제한
보조 테스트 기록은 제조업체가 제공합니다. 장치는 FCC 규칙, Part 15 준수합니다. 장치는
다음 조건에 따라 작동해야 합니다.
1. 유해한 간섭을 일으키지 않아야 합니다.
2. 바람직하지 않은 작동을 유발할 있는 간섭을 포함하여 수신되는 모든 간섭에도 정상적으로
동해야 합니다.
장치의 준수 책임이 있는 측이 명시적으로 허용하지 않은 변경 또는 수정을 가하는 경우 해당 사용
자의 장치 작동 권한이 무효화될 있습니다. 장치는 FCC 규칙, Part 15 의거하여 등급 A 디지
장치 제한 규정을 준수합니다. 이러한 제한은 상업 지역에서 장치를 작동할 유해한 간섭으로부
적절하게 보호하기 위하여 제정되었습니다. 장치는 무선 주파수 에너지를 생성 사용하며
출할 있고 사용 설명서에 따라 설치하고 사용하지 않을 경우 무선 통신에 해로운 간섭을 일으킬
있습니다. 주거 지역에서 장치를 사용하면 해로운 간섭을 일으킬 있으며, 경우 사용자는 자비
들여 간섭 문제를 해결해야 합니다. 다음과 같은 방법으로 간섭 문제를 줄일 있습니다.
1. 장치를 전원에서 분리하여 장치가 간섭의 원인인지 여부를 확인합니다.
2. 장치가 간섭을 받는 장치와 동일한 콘센트에 연결된 경우, 장치를 다른 콘센트에 연결해보십시오.
3. 장치를 간섭을 받는 장치로부터 멀리 떨어트려 놓으십시오.
4. 간섭을 받는 장치의 안테나 위치를 바꿔보십시오.
5. 위의 방법들을 함께 이용해보십시오.
제품 개요
화학적 또는 생물학적 위험 존재. 장비를 공중 위생, 공중 안전, 식음료 제조 또는 가공에 관련한 시행
감시 규정 목적으로 처리공정이나 약품 주입 시스템을 감시하기 위하여 사용하는 경우, 장비에
적용되는 모든 규정을 이해하고 준수하며, 장비가 오작동하는 경우 해당 규정에 따라 충분하고 합당한
메커니즘을 보유하는 것은 사용자의 책임입니다.
화재 위험. 제품은 가연성 액체를 사용할 있게 설계되지 않았습니다.
SD900 휴대용 샘플 채취기는 지정된 간격으로 액체 샘플을 채취하여 또는 용기에 보관합니다.
채취기는 다양한 수성 샘플을 비롯하여 독성 오염 물질 부유 물질 채취 용도로 사용할 있습니
. 샘플 채취기는 여러 리테이너, 또는 용기와 함께 설치할 있습니다. 그림 1 참조하십시오.
156
한글
그림 1 제품 개요
1 소형 베이스 6 컨트롤러 연결
2 표준형 단열 베이스 7 펌프
3 중심부 8 컨트롤러
4 전원 9 액정 센서
5 상단 덮개 10 21 L(5.5 gal) 컨테이너용 표준형 단열 베이스
제품 구성
모든 구성품을 수령했는지 확인하십시오. 그림 2 참조하십시오. 품목이 누락되었거나 손상된 경우
에는 제조업체 또는 판매 담당자에게 즉시 연락하십시오.
그림 2 제품 구성
1 베이스(표준형, 소형 또는 복합형) 6 배터리 충전기(옵션)
2 단일 병으로 채취 시의 부속품( 지지대는 변경
있음)
7 AC 전원공급장치(옵션)
3 중심부(컨트롤러) 8 여과기
4 배터리(옵션) 9 흡입 튜브, 비닐 또는 Teflon 라이닝
5 상단 덮개 10 여러 병으로 채취 시의 부속품( 리테이너는
경될 있음)
한글
157
설치
여러 가지 위험이 존재합니다. 해당 전문가가 문서에 의거하여 작업을 수행해야 합니다.
밀폐 공간 주의사항
폭발 위험 XXX
다음 정보는 사용자가 밀폐 공간에 들어갈 발생할 있는 위험을 이해하도록 돕기 위해 제공되는
것입니다.
1993 4 15 , OSHA CFR 1910.146, 작업 허가 필요 밀폐 공간에 대한 최종 판결이 입법되었
습니다. 표준은 미국 250,000 사이트에 직접적인 영향을 미치며, 밀폐 공간에서의 작업자
안전을 보호하기 위해 제정되었습니다.
밀폐 공간의 정의:
밀폐 공간은 다음 하나 이상의 조건을 가지는 (또는 그럴 가능성이 임박한) 위치 또는 외함입니다.
산소 농도가 19.5% 미만 또는 23.5% 초과 /또는 황화수소(H
2
S) 농도가 10 ppm 초과하는 대기
가스, 증기, 연무, 분진 또는 파이버로 인해 가연성 또는 폭발성이 있는 대기
접촉 또는 흡입 부상, 건강 유해성 또는 사망을 초래할 있는 독성 물질
밀폐 공간은 인간이 점유하도록 설계된 것이 아닙니다. 밀폐 공간은 출입이 제한되며, 알려진 또는
재적인 위험을 가지고 있습니다. 밀폐 공간의 예로는 맨홀, 스택, 파이프, 대형 , 스위치 볼트 기타
유사한 위치가 포함됩니다.
밀폐 공간 /또는 유해 가스, 증기, 연무, 분진 또는 파이버가 존재할 있는 위치에 들어가기 전에
표준 안전 절차를 준수해야 합니다. 밀폐 공간에 들어가기 전에 밀폐 공간 출입과 관련된 모든 절차
찾아 숙지하십시오.
기계 설치
사이트 설치 지침
폭발 위험. 기기는 위험 위치에 설치하도록 승인되지 않았습니다.
설치 장소 평가에 대해서는 지침을 참조하십시오.
샘플 채취기를 좁은 공간에 설치한 경우 모든 안전 수칙을 따르십시오. 밀폐 공간 주의사항
페이지의 158 참조하십시오.
설치 장소의 기온은 명시된 범위를 유지해야 합니다. 사양 페이지의 153 참조하십시오.
샘플 채취기를 평평한 지면에 설치하거나 서스펜션 하니스, 서포트 브래킷 또는 스패너 바로 샘플
채취기를 지탱해야 합니다. 맨홀 안에 샘플 채취기 설치 페이지의 159 적절한 설치 지침 문서를
참조하십시오.
측정 지연 시간을 줄이기 위해 최대한 샘플 소스와 가까워야 합니다. 샘플 채취기 배관 연결
페이지의 161 참조하십시오.
운송 속도 제한 최대 수직 양정 제한에 대한 정보는 사양 페이지의 153 참조하십시오.
158
한글
맨홀 안에 샘플 채취기 설치
맨홀 안에 있는 시료수 위에 샘플 채취기를 설치합니다. 스패너 또는 서포트 브래킷으로 샘플 채취
기를 설치합니다. 맨홀 내부에 스패너 바를 설치합니다. 스패너 바는 양쪽 벽을 미는 힘으로 지지합니
. 서포트 브래킷의 폭은 맨홀 뚜껑과 같습니다. 서포트 브래킷은 뚜껑 바로 밑에 설치하여 지지합니
.제조업체 사이트의 세부 설명서를 참조하십시오.샘플 채취기샘플 채취기.
디스트리뷰터 또는 -보틀(full-bottle) 차단장치(옵션)
디스트리뷰터 또는 -보틀 차단장치는 일반적으로 공장에서 설치됩니다. 설치 디스트리뷰터 또는
-보틀 차단 장치 문서를 참조하십시오.
샘플 채취기 준비
샘플 청소
브러시, , 연성 세제로 샘플 병과 마개를 청소합니다. 병을 깨끗한 물로 씻은 다음 증류수로 헹굽니
.
단일 용기 설치
1 개의 병을 사용하여 하나의 복합 샘플을 채취합니다. 병이 가득 차면 용기 경우 중단 기능이 샘플
프로그램을 중지합니다.
1. 샘플 병을 청소합니다. 샘플 청소 페이지의 159 참조하십시오.
2. 그림 3 나온 대로 샘플 병을 설치합니다.
그림 3 단일 용기 설치
1 10 L(2.5 gal) 폴리에틸렌 (1918) 6 21 L(5.5 gal) 폴리에틸렌 (6494)
2 10 L(2.5 gal) 유리병 (6559) 7 소형 베이스 (8975)
3 지지대 (1502) 8 표준형 단열 베이스 (8976)
4 15 L(4 gal) 폴리에틸렌 (1367) 9 21 L(5.5 gal) 표준형 단열 베이스 (8561)
5 19 L(5 gal) 폴리에틸렌 (6498)
여러 샘플 채취 번째 설치
여러 개의 병을 사용하여 이상의 병에 샘플을 각각 채취합니다. 디스트리뷰터는 위에서
튜브를 움직입니다. 그림 4 나와 있는 것과 같이 병을 샘플 채취기 베이스에 설치합니다. 번째
샘플 (1 ) 샘플 채취기 베이스의 라벨 아래에 설치합니다. 남아 있는 병을 라벨에 나와 있는 숫자
순서대로 설치합니다. 필수 구성 요소의 도표는 그림 5 페이지의 161 참조하십시오.
한글
159
그림 4 1 설치
1 1 위치(소형 베이스) 6 고무줄
2 24 사용 1 위치 7 21 L(5.5 gal) 표준형 단열 베이스
3 8 사용 1 위치 8 표준형 단열 베이스
4 2 또는 4 사용 1 위치 9 소형 베이스
5 리테이너
다중 용기 설치
여러 개의 병을 설치할 경우 디스트리뷰터 암이 샘플 튜브를 위로 이동시킵니다. 지정된 수의
플이 채취되면 샘플 채취가 자동으로 중지합니다.
1. 샘플 병을 청소합니다. 샘플 청소 페이지의 159 참조하십시오.
2. 번째 샘플 (1 ) 샘플 채취기 베이스의 라벨 아래에 설치합니다. 여러 샘플 채취 번째
설치 페이지의 159 참조하십시오.
3. 그림 5 나온 대로 샘플 병을 조립합니다. 병이 8 이상일 경우 번째 병이 시계 방향으로 1
표시기 옆에 오는지 확인합니다.
4. 어셈블리를 시료 채취기에 놓습니다. 병이 8 이상일 경우 와이어를 하단 트레이의 슬롯과
춥니다.
160
한글
그림 5 다중 용기 설치
1 포지셔너/리테이너 (2347) 8 3.8 L(1 gal) 유리병 4 (2216) 또는 3.8 L(1 gal)
리에틸렌 4 (2217)
2 리테이너 (1422) 9 1.9 L(0.5 gal) 유리병 8 (1118) 또는
2.3 L(0.61 gal) 폴리에틸렌 8 (657)
3 리테이너 (2190) 10 1 L(0.26 gal) 폴리에틸렌 24 (737)
4 리테이너 (2189) 11 350 mL(11.83 oz) 유리병 24 (732)
5 950 mL(32.12 oz) 유리병 8 (2348) 12 소형 베이스 (8975)
6 575 mL(19.44 oz) 폴리에틸렌 24 (1369) 13 표준형 베이스 (8976)
7 3.8 L(1 gal) 유리병 2 (2214) 또는 3.8 L(1 gal)
리에틸렌 2 (2215)
샘플 채취기 배관 연결
대표 샘플이 채취되도록 (표면 또는 바닥 가까이가 아니라) 시료 흐름의 중앙에 여과기를 설치합니다.
그림 6 참조하십시오. 흡입 튜브 설치 그림 7 참조하십시오.
1. 그림 7 나온 대로 시료 채취기에 튜브를 연결합니다.
참고
: Teflon
라이닝
튜브를
사용하는
경우
연결
키트
2186
사용합니다
.
2. 흡입 튜브 여과기를 물이 요동치고 섞이는 샘플 소스의 본류에 설치합니다.
흡입 튜브는 가능한 짧게 만듭니다.
샘플 사이에 흡입 튜브가 완전히 배수되도록 튜브를 최대한 수직으로 유지합니다.
참고
:
수직이
불가능하거나
튜브가
가압되는
경우
액정
센서를
사용하지
마십시오
.
시료
볼륨을
수동으로
정하십시오
. 시료 볼륨을 수동으로 교정
페이지의
171
참조하십시오
.
흡입 투브가 집히지 않았는지 확인합니다.
한글
161
그림 6 기기 설치
1 여과기 3 흡입 튜브
2 수직 양정
그림 7 흡입 튜브 설치
전기 설치
컨트롤러 연결
전기쇼크 위험. 외부 연결된 장비는 해당하는 국가 안전 표준에 따라 평가를 마쳐야 합니다.
그림 8 에는 컨트롤러에 연결할 있는 커넥터가 나와 있습니다.
162
한글
그림 8 컨트롤러 커넥터
1 보조 장치 4 SDI-12 장치 옵션
2 전원공급장치 5 디스트리뷰터/용기 경우 중단
3 시리얼 통신
유량계 연결
유량계를 컨트롤러에 연결하여 샘플 유량이 지정된 값을 상회 또는 하위하면 샘플 채취기를 시작하거
중지합니다.
준비 항목:
Sigma 유량계용 다목적 케이블 (또는 모델 980 유량계용 980 하프 케이블)
추가 연결용 분배기 (옵션). 2 이상의 분배기를 직렬로 연결할 있습니다.
1. 케이블의 한쪽 끝을 유량계에 연결합니다. 모델 980 유량계의 경우 모델 980 유량계 사용 설명서
참조하십시오.
2. 케이블의 다른 끝을 컨트롤러의 보조 장치 커넥터에 연결합니다.
참고
:
유량계가
6
케이블을
사용할
경우
6
-7
어댑터
케이블을
사용하십시오
.
PC 또는 통신 네트워크 연결
데이터를 전송하거나 샘플 채취기를 구성하려면 PC 또는 통신 네트워크를 컨트롤러에 연결합니다.
준비 항목:
시리얼 케이블, 7 RS232 - DB-9 (그림 9 참조). 연결: B - 5 (신호 접지), D - 3 (RCD), F - 2 (TXD),
G (접지)
1. 시리얼 케이블의 한쪽 끝을 통신 장치 또는 네트워크에 연결합니다.
2. 케이블의 다른 끝을 컨트롤러의 시리얼 통신 커넥터에 연결합니다.
그림 9 시리얼 통신 케이블
한글
163
샘플 채취기 전원 연결
전기 쇼크 위험. 기기를 실외 또는 습기있는 장소에서 사용하는 경우, 접지 결함 회로 인터럽트
(GFCI/GFI) 장치를 사용하여 기기를 메인 전원에 연결 합니다.
화재 위험. 전원 라인에 15 A 용량의 회로 차단기를 설치하십시오. 회로 차단기는 장비 근처에 설치했을
국부 전원 차단 역할을 합니다.
전기쇼크 위험 보호 접지(PE) 연결이 필요합니다.
전기쇼크 위험 해당 위치에서 전원을 쉽게 차단할 있어야 합니다.
감전 위험. 펌프 사이클 사이의 시간이 너무 짧으면 전원 공급기가 과열될 있습니다.
펌프 사이클 시간 때문에 전원공급장치가 과열되지 않아야 합니다. 펌프 사이클 시간 계산 페이지의 165 참조하
십시오.
샘플 채취기에 배터리 또는 AC 전원공급장치를 연결합니다. 아래의 단계별 그림 설명을 참조하십시
.
164
한글
펌프 사이클 시간 계산
전원공급장치가 과열되지 않도록 샘플 사이클 사이에 펌프가 최소 기간 동안 작동을 중지해야 합니다.
1. 샘플 사이클 동안 펌프가 연속 작동할 시간을 계산합니다. 다음 단계를 모두 포함합니다. 정화
, 흡입 튜브 린스, 샘플, 샘플 재시도 정화 .
2. 그림 10 에서 샘플 사이클 사이에 펌프가 작동을 중지해야 하는 최소 시간을 찾습니다.
3. 샘플 채취기 프로그램의 페이싱 또는 시간 주기가 샘플 사이클 사이에 최소 시간 동안 펌프를 작동
중지 상태로 유지하는지 확인합니다.
그림 10 50 °C (122 ºF) 펌프 작동 중지 횟수
1 펌프가 작동 중지해야 하는 시간() 3 : 펌프가 5 분간 연속 작동할 경우 펌프는 15 분간
작동을 중지해야 합니다.
2 펌프가 연속 작동하는 시간()
전원 켜기
전원 버튼을 눌러 전원을 켜거나 끄십시오. 전원공급장치(AC 전원 또는 배터리) 올바르게 설치됐는
확인하십시오.
사용자 인터페이스 탐색
컨트롤러 키패드는 그림 11 나와 있습니다. 샘플 채취 프로그램을 설정하거나 컨트롤러 설정을
성하거나 진단 시험을 실시하려면 MENU(메뉴) 키를 사용합니다. 메뉴를 스크롤하고, 항목을 선택하
, 값을 입력하려면 화살표, ENTER(엔터) BACK(뒤로) 키를 사용합니다. 많은 옵션을 사용할
있는 경우 디스플레이 화면의 하단 또는 측면에 화살표 키가 표시됩니다.
한글
165
그림 11 컨트롤러 키패드
1 POWER(전원) 5 메뉴 9 상태
2 VOLUME CALIBRATION(볼륨
교정)
6 ENTER 10 RUN/HALT PROGRAM(프로그
실행/중단)
3 정지 7 뒤로 11 MANUAL OPERATION(수동
)
4 LED 8 화살표
작동
시스템 설정 구성
1. MENU(메뉴) 누릅니다.
2. SYSTEM SETUP(시스템 설정)으로 이동합니다.
3. 해당 옵션을 변경합니다.
옵션 설명
시간/날짜 설정 시간 (24 시간 형식) 날짜를 설정합니다.
통신 시리얼 포트의 전송 속도 (19200, 38400, 57600 또는 115200) 프로토콜 (Modbus RTU
또는 ASCII) 선택합니다.
기본 설정 샘플 채취기 베이스를 선택합니다(전천후 냉장형, 냉장형, 휴대용 표준형 또는 휴대용
).
언어 컨트롤러 언어를 선택합니다.
액정 센서 교정 액정 센서를 교정하거나 기본값으로 설정합니다.
명암비 설정 LCD 화면의 명암비를 조정합니다.
펌프 듀티 사이클 펌프 듀티 사이클을 변경합니다. 범위: 50~100%(기본값: 100%)
암호 설정 프로그램 설정 시스템 설정 메뉴에 대한 액세스를 제한하는 암호를 설정합니다.
ENABLE(사용)>YES() 선택하여 암호를 설정하거나 ENABLE(사용)>NO(아니요)
선택하여 현재 암호를 활성화합니다. 암호를 최초로 설정하는 경우 현재 암호로 900900
사용하십시오.
166
한글
옵션 설명
튜브 수명 펌프 튜브 교체 시기를 안내하는 알림 메시지를 설정합니다. ENABLE(사용)>RESET
CYCLES(사이클 재설정)으로 이동하여 펌프 사이클 카운트를 0 으로 재설정합니다.
ENABLE(사용)>CYCLE LIMIT(사이클 한계) 이동하여 알림 메시지에 사용할 펌프 사이클
수를 변경합니다. ENABLE(사용)>STATUS(상태) 이동하여 펌프 사이클 카운트의 현재
수치를 표시합니다.
SDI–12 (옵션) SDI–12 장치의 작동을 구성합니다. 옵션은 장치가 발견될 때에만 표시됩니다.
샘플 채취 프로그램 설정
샘플 채취 프로그램에는 자동으로 샘플을 채취하는 필요한 모든 매개변수가 포함됩니다. 정기적으
또는 샘플 채취 프로그램이 완료되면 샘플을 채취합니다.
1. MENU(메뉴) 누릅니다.
2. PROGRAM SETUP(프로그램 설정)>MODIFY ALL(모두 수정)으로 이동합니다. 번째 화면이
시됩니다.
3. 매개변수의 값을 선택하거나 입력합니다. 왼쪽 또는 오른쪽으로 이동하려면 왼쪽/오른쪽 화살
키를 사용합니다. 값을 변경하려면 위쪽/아래쪽 화살표 키를 사용합니다. 필요할 경우 화살표
사용하여 단위 (gal 또는 mL) 변경합니다.
옵션 설명
용기 용기 : 샘플 채취기에 설치된 병의 개수(1, 2, 4, 8 또는 24).
용기 경우 중단: 사용/사용 . 사용으로 설정할 경우 컨트롤러가 병이 넘치는지 여부를 점검
합니다.
용량: 병의 용량. 범위: 50~65000 mL(0.01~17.17 gal)
흡입
튜브 길이: 여과기에서 액정 센서까지의 흡입 튜브 길이. 범위: 100–3000 cm 또는 3–99 ft. 정확한
샘플 볼륨을 얻으려면 정확한 길이가 필요합니다.
흡입 튜브 유형: 흡입 튜브의 직경 재질 (¼ in. 비닐, 3/8 in. 비닐 또는 3/8 in. Teflon).
프로그램
지연
사용/사용 : 사용으로 설정할 경우 지정된 시간 또는 카운트 수에서 샘플 채취 프로그램이 시작
합니다. 날짜 시간: 프로그램이 시작하는 날짜 시간 (24 시간 형식). 카운트: 프로그램이 시작
유량계 카운트 또는 펄스 (1-9999 카운트). 나중에 샘플 페이싱을 시간으로 설정할 경우 프로
그램 지연이 비활성화됩니다.
샘플
이싱/
시간 기반 - 시간 기반 수집 페이지의 168 참조하십시오.
참고
: CTVV
사용되는
경우
:
액정
센서가
항상
활성화됩니다
.
샘플
분산이
합성됩니다
.
실행
모드가
계속되지
않습니다
.
번째
샘플이
간격
이후
수집됩니다
.
유량
소스
입력은
항상
4–20 mA
입니다
.
유량 기반 - 유량 기반 수집 페이지의 169 참조하십시오.
샘플
샘플을 모든 병으로 분배>
샘플이 모든 병으로 분배됩니다. 마지막 샘플 종료: 프로그램이 마지막 샘플 종료합니다
(1–999). 연속: 프로그램이 수동으로 중지될 때까지 계속됩니다.
샘플을 모든 병으로 분배>아니요
샘플이 지정된 그룹의 병으로 분배됩니다. . 샘플/용기: 병에 채취될 샘플 (1–999). 용기/샘플:
동일한 샘플이 채취될 병의 개수.
액정
사용 또는 사용 . 사용 함으로 설정할 경우 샘플 볼륨을 시간 기반으로 교정해야 합니다.
시료
샘플의 볼륨 (단위: mL (1 gal = 3785.4 mL)). 용기/샘플 모드를 사용할 경우 병에는 전체 샘플
볼륨이 들어갑니다 (10–10,000 mL). 샘플 볼륨이 용량을 초과하지 않는지 확인하십시오. 샘플
볼륨은 10 mL 단위로 반올림합니다.
한글
167
옵션 설명
흡입
린스
샘플 채취 흡입 튜브 린스 횟수 (0–3).
샘플
시도
샘플 채취가 실패할 경우 샘플 채취 시도 횟수 (0–3).
사이트
ID
샘플 채취 위치의 이름 (최대 12 ). 프로그램이 저장되면 사이트 ID 사전 설정 이름으로 사용됩
니다.
고급
채취
고급 샘플 채취 메뉴로 이동합니다. 필요할 경우 고급 샘플 채취 메뉴에서 옵션을 구성합니다.
프로그램
실행
기본 샘플 프로그램 생성이 완료됩니다. 샘플 채취기가 프로그램을 시작 또는 취소하라는 메시지를
표시합니다.
시간 기반 수집
1. 옵션의 값을 선택하거나 입력합니다.
옵션 설명
속도 간격 샘플을 일정한 시간 간격(단위: 시간 단위)으로 수집합니다 (0:01–999:00).
번째 샘플
프로그램은 즉시 시작하거나 번째 간격 이후에 시작합니다.
변화 가능한
샘플 볼륨의 변화 여부를 지정합니다(유량 기반). 변화할 경우 변화 가능한 볼륨 사용
페이지의 168 참조합니다. 변화하지 않을 경우 변화 가능한 볼륨을 비사용
페이지의 169 참조합니다.
변화 가능한 볼륨 사용
샘플 속도는 일정 시간 동안 변화 가능 볼륨(CTVV) 입니다.
1. 옵션의 값을 선택하거나 입력합니다.
옵션 설명
단위 선택 사용할 유량 단위를 선택합니다.
MAP 4 mA 입력 4 mA 입력에 일치하는 유속을 입력합니다.
MAP 20 mA 입력 20 mA 입력에 일치하는 유속을 입력합니다.
평균 유속 사이트의 평균 유속을 지정합니다(1–999,999). 단위는 위에서 선택한 단위로 지정됩니다.
원하는 시간 샘플 프로그램 도중 수집하고자 하는 볼륨을 지정합니다(10–10,000 ml).
168
한글
옵션 설명
수집 기간 수집 기간을 시간과 분으로 지정합니다(0:01–999:00).
최소 샘플량 유량이 최소 볼륨을 수집하기에 부족할 경우 샘플이 생략됩니다(10–10,000 ml).
변화 가능한 볼륨을 비사용
1. 옵션의 값을 선택하거나 입력합니다.
옵션 설명
번째 샘플 수집 프로그램을 즉시 시작할 , 번째 간격 이후에 시작할 선택합니다.
유량 기반 수집
유량 기반 샘플의 흐름 소스를 다음과 같이 지정합니다.
카운트 - 유량 소스에 카운트 사용 페이지의 169 참조하십시오.
4–20 mA 입력 - 유량 소스에 4–20 mA 입력 페이지의 169 참조하십시오.
유량 소스에 카운트 사용
1. 옵션의 값을 선택하거나 입력합니다.
옵션 설명
간격마다 샘플 수집 일정 간격(단위: 카운트)으로 샘플을 채취합니다 (1–9999 카운트).
중지 시간 유량이 너무 적을 경우 샘플을 채취합니다.
시간 샘플 최대 시간 (0:01–999:00). 샘플이 수집되면 타이머가 다시 시작합니다.
번째 샘플 수집 프로그램이 즉시 시작하거나 번째 간격 이후에 시작합니다.
유량 소스에 4–20 mA 입력
1. 옵션의 값을 선택하거나 입력합니다.
옵션 설명
단위 선택 사용할 유량 단위를 선택합니다.
MAP 4 mA 입력 4 mA 일치하는 유속을 입력합니다.
MAP 20 mA 입력 20 mA 일치하는 유속을 입력합니다.
유량마다 샘플
샘플을 수집할 유량을 입력합니다.
중단 시간 유량 볼륨이 비정상적으로 낮을 샘플을 수집하려면 선택합니다.
시간 샘플 최대 시간 (0:01–999:00). 샘플이 수집되면 타이머가 다시 시작합니다.
시료 볼륨 샘플 수집할 볼륨을 입력합니다. /샘플 모드를 사용할 경우 병에는 전체 샘플
볼륨이 들어갑니다 (100–10,000 mL). 변화 가능한 볼륨이 선택된 경우 옵션은 생략
됩니다.
기본 설정 복원
복원 옵션은 모든 프로그램 매개변수를 기본 설정으로 복원합니다.
1. MENU(메뉴) 누릅니다.
2. PROGRAM SETUP(프로그램 설정)>RESTORE(복원) 이동합니다.
3. YES() 선택합니다.
한글
169
프로그램 시작 또는 중지
RUN/HALT PROGRAM(프로그램 실행/중단) 키를 사용하여 프로그램을 시작하거나 중지합니다. 최대
12 개의 시작/중지 시간/날짜를 사용자 정의할 있습니다. 프로그램 또는 시스템 설정을 변경하려면
프로그램을 중지해야 합니다.
1. RUN/HALT PROGRAM(프로그램 실행/중단) 누릅니다.
2. 옵션 하나를 선택합니다.
옵션 설명
시작 현재 로드된 프로그램을 시작합니다. 상태가 RUNNING(실행 )으로 바뀝니다.
중단 프로그램을 일시적으로 중지합니다. 상태가 HALTED(중단됨)으로 바뀝니다.
재개 프로그램이 중단된 지점에서 재개됩니다. 상태가 RUNNING(실행 )으로 바뀝니다.
처음부터 시작 프로그램이 처음부터 시작합니다. 상태가 RUNNING(실행 )으로 바뀝니다.
프로그램 종료 현재 실행 중인 프로그램을 종료합니다. 상태가 COMPLETE(완료) 바뀝니다.
수동 작동
수동 작동을 사용하여 그랩 샘플을 채취하거나 디스트리뷰터 암을 이동하거나 펌프를 작동합니다.
세한 내용은 설명서의 확장판을 참조하십시오.
볼륨 교정
샘플 볼륨은 액정 센서를 통해 자동으로, 또는 수동으로 교정할 있습니다. 액정 센서를 사용하여
교정을 실시할 경우 센서가 모든 프로그래밍된 볼륨을 정확하게 측정하도록 조정됩니다. 액정 센서
사용 함으로 설정한 경우에는 샘플, 린스 우수용 볼륨을 수동으로 교정해야 합니다.
일반 용도에서는 수돗물을 사용하여 볼륨 교정을 실시할 있습니다. 샘플 조성이 일반적인 샘플
현저히 다른 경우 샘플 채취기를 채취할 액체로 교정하십시오. 교정을 실시하려면 모든 프로그램을
중지해야 합니다.
액정 센서 교정
액정 센서의 정확도는 샘플로 채취되는 액체의 종류에 따라 다릅니다. 예를 들어 매우 혼탁한 샘플의
볼륨은 맑은 물만큼 정확하지 않을 있습니다. 액정 센서는 샘플로 교정하여 정확도를 개선할
있습니다.
1. 흡입 튜브를 샘플에 넣습니다.
2. MENU(메뉴) 누릅니다.
3. PROGRAM SETUP(프로그램 설정)>MODIFY SELECTED(선택 수정)>LIQUID SENSOR(액정
)>ENABLE(사용)으로 이동하여 액정 센서가 사용으로 설정되어 있는지 확인합니다.
4. SYSTEM SETUP(시스템 설정)>LIQ SENSOR CAL(액정 센서 교정)으로 이동합니다.
5. PERFORM CAL(교정 실시) 선택합니다.
6. START(시작) 선택합니다. 펌프가 역방향으로 작동하여 흡입 튜브를 정화합니다. 그런 다음
프가 정방향으로 작동합니다.
7. 흡입 튜브 안의 샘플 흐름을 관찰하여 샘플이 액정 센서를 통과하는지 확인합니다.
8. 샘플이 액정 센서를 통과하고 기포가 없으면 STOP(중지) 누릅니다. 흡입 튜브가 자동으로 정화
됩니다. 액정 센서 교정이 완료되었습니다.
9. DONE(완료) 선택하여 종료하거나 REPEAT(반복) 선택하여 교정을 다시 실시합니다.
기본 교정 복원
액정 센서의 볼륨 교정을 출고 기본 설정으로 복원할 있습니다.
170
한글
1. PROGRAM SETUP(프로그램 설정)>MODIFY SELECTED(선택 수정)>LIQUID SENSOR(액정
)>ENABLE(사용) 선택하여 액정 센서를 사용으로 설정합니다.
2. VOLUME CALIBRATION(볼륨 교정) 누르고 RESET CAL(교정 재설정) 선택합니다. 볼륨
정이 기본 설정으로 설정됩니다. “값이 0 으로 재설정됨메시지가 표시됩니다.
액정 센서를 사용하여 샘플 볼륨 교정
액정 센서를 사용으로 설정한 상태에서 샘플 볼륨을 교정하여 샘플 볼륨을 약간 조정합니다.
1. PROGRAM SETUP(프로그램 설정)>MODIFY SELECTED(선택 수정)>LIQUID SENSOR(액정
)>ENABLE(사용)으로 이동하여 액정 센서가 사용으로 설정되어 있는지 확인합니다.
2. 흡입 튜브를 샘플 또는 수돗물에 넣습니다.
참고
:
샘플을
사용하면
최상의
정확도를
얻을
있습니다
.
3. 펌프의 출구 측에서 나온 튜브를 메스 실린더에 넣습니다.
4. VOLUME CALIBRATION(볼륨 교정) 누르고 CALIBRATION(교정) 선택합니다.
5. START(시작) 선택합니다. 펌프가 역방향으로 작동하여 흡입 튜브를 정화합니다. 그런 다음
프가 정방향으로 작동하여 샘플 채취 프로그램에 지정된 볼륨을 채취합니다. 펌프가 역방향으로
작동하여 흡입 튜브를 정화합니다.
6. 샘플이 채취되면 DONE(완료) 선택합니다. 메스 실린더 안의 샘플 볼륨과 프로그램 설정 메뉴에
입력된 샘플 볼륨을 비교합니다.
7. 채취된 볼륨이 현재 프로그램의 샘플 볼륨과 다를 경우 실제로 채취된 볼륨을 입력합니다. ENTER
누릅니다. 센서가 모든 프로그래밍된 볼륨을 정확하게 측정하도록 조정됩니다.
8. START(시작) 선택하여 샘플 채취 프로그램을 시작하거나 CANCEL(취소) 선택하여 볼륨 교정
메뉴를 종료합니다.
시료 볼륨을 수동으로 교정
액정 센서를 사용 함으로 설정한 경우 모든 볼륨을 시간 기반으로 수동으로 교정해야 합니다. 샘플
볼륨이 현재 프로그램에 지정된 볼륨에 대해 교정됩니다. 프로그램에서 샘플 볼륨을 변경할 경우
볼륨에 대해 샘플 볼륨을 다시 교정해야 합니다.
1. PROGRAM SETUP(프로그램 설정)>MODIFY SELECTED(선택 수정)>LIQUID SENSOR(액정
)>DISABLE(사용 ) 선택하여 액정 센서를 사용 함으로 설정합니다.
2. 흡입 튜브를 샘플 또는 수돗물에 넣습니다.
3. 펌프의 출구 측에서 나온 튜브를 메스 실린더에 넣습니다.
4. VOLUME CALIBRATION(볼륨 교정) 누릅니다. 2 이상의 볼륨이 표시되면 교정할 볼륨 1
선택합니다.
5. START(시작) 선택합니다. 펌프가 역방향으로 작동하여 흡입 튜브를 정화합니다. 그런 다음
프가 정방향으로 작동하여 샘플을 채취하기 시작합니다.
6. 선택된 볼륨 또는 린스 위치에서 펌프 작동을 중지합니다.
샘플 볼륨: 메인 프로그램에서 지정된 볼륨이 채취되면 STOP(중지) 누릅니다.
린스: 흡입 튜브 안의 액체가 액체 검출기에 도달하면 STOP(중지) 누릅니다.
우수 볼륨: 우수 프로그램에서 지정된 볼륨이 채취되면 STOP(중지) 누릅니다.
7. DONE(완료) 선택하여 다음 볼륨 교정으로 이동하거나 REPEAT(반복) 선택하여 교정을 다시
실시합니다.
8. 모든 볼륨을 교정했으면 DONE(완료) 선택합니다.
9. 배출 튜브를 샘플 채취기의 튜브 피팅에 연결합니다.
10. START(시작) 선택하여 샘플 채취 프로그램을 시작하거나 CANCEL(취소) 선택하여 종료합니
.
한글
171
샘플 볼륨 검증
샘플 볼륨이 정확한지 확인하려면 그랩 샘플을 채취합니다. 볼륨을 점검하기 위해 교정으로 돌아가지
마십시오. 교정 시작 볼륨 보상이 0 으로 재설정됩니다.
1. MANUAL OPERATION(수동 작동) 누릅니다.
2. GRAB SAMPLE(그랩 샘플) 선택합니다.
3. 흡입 튜브를 샘플에 넣습니다.
4. 펌프의 출구 측에서 나온 튜브를 메스 실린더에 넣습니다.
5. 액정 센서를 사용으로 설정한 경우 검증할 볼륨을 입력합니다.
6. 액정 센서를 사용 함으로 설정한 경우 프로그램에서 지정된 볼륨을 입력합니다.
7. ENTER 누릅니다. 펌프 사이클이 시작합니다 (정화-샘플-정화).
8. 메스 실린더에 채취된 볼륨을 예상 볼륨과 비교합니다. 채취된 볼륨이 정확하지 않으면 볼륨 교정
다시 실시합니다.
유지관리
여러 가지 위험이 존재합니다. 해당 전문가가 문서에 의거하여 작업을 수행해야 합니다.
전기쇼크 위험 유지관리 또는 정비 작업을 수행하기 전에 기기의 전원을 차단하십시오.
생물학적 위험 노출. 샘플 샘플 채취기 성분과 접촉할 때는 안전 취급 규정을 따르십시오.
여러 가지 위험이 존재합니다. 기술자는 유지관리 절차 장치가 안전하고 올바르게 작동하는지 확인해
합니다.
기기 세척
화재 위험. 가연성 세정제를 사용하여 기기를 세척하지 마십시오.
컨트롤러 히터를 어떠한 종류의 액체로도 청소하지 마십시오.
컨트롤러 펌프를 물로는 청소할 없을 경우 컨트롤러를 분리하여 샘플 채취기와 떨어뜨려 놓습니
. 컨트롤러 펌프가 마르도록 충분히 기다린 부품을 다시 설치하고 사용을 재개하십시오.
다음과 같이 샘플 채취기를 청소합니다.
샘플 채취기 케이스 - 내부 외부 표면은 젖은 헝겁과 중성 세제를 사용하여 청소하십시오. 마모성
클리너 또는 솔벤트를 사용하지 마십시오.
샘플 - 병은 브러시와 중성 세제를 섞은 물을 사용하여 청소하십시오. 깨끗한 물로 헹군 다음
류수로 헹구십시오. 필요시 고압 멸균 처리기를 사용하여 유리병을 청소하십시오.
172
한글
건조제 교체
컨트롤러 건조제 카트리지는 습기를 흡수하고 부식을 방지합니다. 창을 통해 건조제 색상을 모니터
링합니다 (그림 12). 건조제는 노란색입니다. 녹색으로 변하면 건조제를 교체합니다.
1. 나사를 풀고 건조제 카트리지를 분리합니다(그림 12).
2. 플러그를 분리하고 사용한 건조제를 버립니다.
3. 건조제 튜브에 건조제를 채웁니다.
4. 플러그를 설치합니다.
5. 오링에 그리스를 바릅니다.
6. 건조제 튜브를 컨트롤러에 설치합니다.
그림 12 건조제 카트리지
1 플러그 3 오링
2 건조제 튜브 4 건조제
펌프 유지관리
압착 위험. 유지관리 또는 정비 작업을 수행하기 전에 기기의 전원을 차단하십시오.
펌프 배관 교체
제조업체에서 제공하지 않은 배관을 사용할 경우 기계 부품이 과도하게 마모될 있고 그리고/또는 펌프 성능이
저하될 있습니다.
롤러가 펌프 배관과 마찰되는 부위에서 배관의 마모 여부를 검사합니다. 알림 메시지가 표시되거나
관에서 마모 징후가 보이면 배관을 교체하십시오.
사전 필요 조건:
펌프 배관프리컷 또는 벌크 4.6 m 또는 15.2 m (15 ft 또는 50 ft)
1. 컨트롤러 전원을 분리합니다.
2. 벌크 배관을 사용하는 경우 배관을 절단하고 정렬 도트를 추가합니다. 제조업체 사이트의 세부
설명서를 참조하십시오.
3. 아래의 그림 단계에 따라 펌프 배관을 분리합니다.
4. 펌프 하우징 내부와 롤러에서 실리콘 잔류물을 제거합니다.
5. 아래의 그림 단계에 따라 펌프 배관을 설치합니다.
한글
173
174
한글
로터 청소
연성 세제로 로터, 펌프 튜브 트랙 펌프 하우징을 청소합니다. 펌프 배관 교체 페이지의 173 아래
그림 단계를 참조하십시오.
디스트리뷰터 튜브 교체
디스트리뷰터 암은 다중 용기 샘플 채취 위로 이동합니다. 튜브가 마모된 경우 디스트리뷰터
암의 튜브를 교체합니다. 디스트리뷰터 디스트리뷰터 암에 적합한 튜브를 사용해야 합니다.
참고
:
디스트리뷰터와
펌프
튜브는
서로
다릅니다
.
디스트리뷰터
어셈블리에
설치된
펌프
튜브로
인해
디스트리뷰터
손상될
있습니다
.
또한
디스트리뷰터
암이
제대로
이동하지
못하여
샘플이
손실될
있습니다
.
1. 디스트리뷰터 암과 중심부 천정에서 튜브를 제거합니다.
2. 디스트리뷰터 암에 튜브를 끼웁니다. 튜브는 그림 13 같이 규격에 맞춰야 합니다.
3. 튜브의 다른 끝을 중심부 천정에 있는 피팅에 끼웁니다.
4. 디스트리뷰터 진단 시험을 실시하여 올바로 작동하는지 확인합니다.
그림 13 디스트리뷰터 어셈블리
1 노즐 3 디스트리뷰터 모터 5 디스트리뷰터 튜브
2 디스트리뷰터 4 샤프트
퓨즈 교체
퓨즈는 사용자가 취급할 있는 품목이 아닙니다. 기술 지원부에 문의하십시오. 퓨즈 단절은 샘플
취기에 문제가 생겼다는 징후이며, 경우에는 서비스를 요청해야 합니다.
한글
175
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
Tel. (970) 669-3050
(800) 227-4224 (U.S.A. only)
Fax (970) 669-2932
www.hach.com
HACH LANGE GMBH
Willstätterstraße 11
D-40549 Düsseldorf, Germany
Tel. +49 (0) 2 11 52 88-320
Fax +49 (0) 2 11 52 88-210
www.hach-lange.de
HACH LANGE Sàrl
6, route de Compois
1222 Vésenaz
SWITZERLAND
Tel. +41 22 594 6400
Fax +41 22 594 6499
©
Hach Company/Hach Lange GmbH, 2013.
All rights reserved. Printed in U.S.A.

Transcripción de documentos

DOC026.97.80437 SD900 Portable Sampler 02/2015, Edition 2 Basic User Manual Manuel d'utilisation de base Manual básico del usuario Manual Básico do Usuário 基本用户手册 基本取扱説明書 기본 사용 설명서 English .............................................................................................................................. 3 Français ......................................................................................................................... 28 Español .......................................................................................................................... 54 Português ...................................................................................................................... 80 中文 ............................................................................................................................... 106 日本語 ........................................................................................................................... 129 한글 ............................................................................................................................... 153 2 Table of contents Specifications on page 3 Operation on page 17 General information on page 5 Maintenance on page 23 Installation on page 8 Additional information Additional information is available on the manufacturer's website. Specifications Specifications are subject to change without notice. SD900 Portable Sampler Specification Standard base Compact base Composite base Dimensions Diameter: 50.5 cm (19.9 in.) Diameter: 44.1 cm (17.4 in.) Height: 69.4 cm (27.3 in.) Height: 61 cm (24 in.) Diameter: 50.28 cm (19.8 in.) Height: 79.75 cm (31.4 in.) Weight 15 kg (35.6 lb) with 1-L polyethylene bottles (24x) 14.8 kg (32.6 lb) with 10-L (2.5 gal) polyethylene container (1x) 12.2 kg (27 lb) with 575-mL (19.44 oz) polyethylene bottles (24x) 15 kg (36 lb) with 950mL (32.12 oz) glass bottles (12x) 12.9 kg (28.3 lb) with 10-L (2.5 gal) polyethylene container (1x) Enclosure Impact-resistant ABS, 3-section construction; double-walled base with 2.54 cm (1 in.) insulation—direct bottle contact with ice. Sample temperature 0–60 °C (32–140 °F) Strainers 316 stainless steel in standard size, high velocity or low profile for shallow depth applications and Teflon® or 316 stainless steel in standard size Sample intake tubing 9.5 mm (3/8 in.) I.D. vinyl or Teflon-lined polyethylene Sample bottle capacity 1-L (0.26 gal) polyethylene and/or 350-mL (11.83 oz) glass bottles (24x) 2.3-L (0.6 gal) polyethylene and/or 1.9-L (0.5 gal) glass bottles (8x) 3.8-L (1 gal) polyethylene and/or 3.8-L (1 gal) glass bottles (4x) 3.8-L (1 gal) polyethylene and/or 3.8-L (1 gal) glass bottles (2x) 575-mL (19.44 oz) polyethylene bottles (24x) 21-L (5.5 gal) polyethylene bottle (1x) 950-mL (32.12 oz) glass bottles (8x) 10-L (2.5 gal) polyethylene bottle (1x) 10-L (2.5 gal) glass bottle (1x) 21-L (5.5 gal) or 15-L (4 gal) polyethylene composite container or 20-L (5.25 gal) polyethylene or 10-L (2.5 gal) polyethylene or 10-L (2.5 gal) glass (1x) English 3 SD controller Specification Details Dimensions (W x H x D) 29.2 x 17.1 x 26.4 cm (11½ x 6¾ x 103/8 in.) Weight 4.2 kg (9.26 lb) Enclosure PC/ABS blend, NEMA 4X, 6, IP 67 Power requirements 15 VDC supplied by a 8754500 power supply; 15 VDC supplied by an integral power supply Overload protection 7 A, DC line fuse for the pump Operating temperature 0 to 50 °C (32 to 122 °F) Storage temperature –30 to 60 °C (–22 to 140 °F) Storage/operating humidity 100% condensing Pump Peristaltic high speed, with spring-mounted Nylatron rollers Pump enclosure IP37 Pump tubing 9.5 mm ID x 15.9 OD mm (3/8 in. ID x 5/8 in. OD) silicone Pump tubing life 20,000 sample cycles with: 1 L (0.3 gal) sample volume, 1 rinse, 6 minute pacing interval, 4.9 m (16 ft) of 3/8 in. intake tube, 4.6 m (15 ft) of vertical lift, 21 °C (70 °F) sample temperature Vertical sample lift Maximum 8.5 m (28 ft) for: 8.8 m (29 ft) of 3/8-in. vinyl intake tube at sea level at 20–25 °C (68–77 °F) Pump flow rate 4.8 L/min (1.25 gpm) at 1 m (3 ft) vertical lift with 3/8-in. intake tube typical Sample volume Programmable in 10-mL (0.34 oz) increments from 10 to 10,000 mL (3.38 oz to 2.6 gal) Sample volume repeatability (typical) ±5% of 200 mL sample volume using uncalibrated liquid detect with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of 3/8-in. vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m (5000 ft) elevation Sample volume accuracy (typical) ±10% of 200 mL sample volume using uncalibrated liquid detect with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of 3/8-in. vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m (5000 ft) elevation Sampling modes Pacing: Time-fixed, flow-fixed, time-variable, flow-variable, constant time variable volume (CTVV). Refer to Constant Time Variable Volume (CTVV) sampling in the expanded version of this manual. Distribution: Single bottle composite, multi-bottle composite, multi-bottle discrete, bottles per sample, samples per bottle, combination of bottles per sample, samples per bottle Run modes Continuous or non-continuous with user-entered number of samples Multiple programs Stores up to three sampling programs Transfer velocity (typical) 0.9 m/s (2.9 ft/s) with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of 3/8-in. vinyl intake tubing, 21 °C (70 °F) and 1524 m (5000 ft) elevation Liquid sensor Ultrasonic. Body: Ultem NSF ANSI standard 51 approved, USP Class VI compliant Sample history Up to 510 records Air purge Air purged automatically before and after each sample. The duration automatically compensates for varying intake tube lengths. Event log 510 records 4 English ® Specification Details Connections Power, auxiliary, serial communications, distributor, SDI-12 Wetted materials Stainless steel, polyethylene, Teflon, Ultem, silicone Communications RS232, Modbus, SDI-12 Warranty 1 year General information In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website. Safety information NOTICE The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction. Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment. Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual. Use of hazard information DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNING Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury. CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury. NOTICE Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis. English 5 Precautionary labels Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement. This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information. This symbol indicates a potential pinch hazard. Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user. Certification Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A: Supporting test records reside with the manufacturer. This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations. Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences. FCC Part 15, Class "A" Limits Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: 1. The equipment may not cause harmful interference. 2. The equipment must accept any interference received, including interference that may cause undesired operation. Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems: 1. Disconnect the equipment from its power source to verify that it is or is not the source of the interference. 2. If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet. 3. Move the equipment away from the device receiving the interference. 4. Reposition the receiving antenna for the device receiving the interference. 5. Try combinations of the above. 6 English Product overview DANGER Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or chemical feed system for which there are regulatory limits and monitoring requirements related to public health, public safety, food or beverage manufacture or processing, it is the responsibility of the user of this instrument to know and abide by any applicable regulation and to have sufficient and appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction of the instrument. CAUTION Fire hazard. This product is not designed for use with flammable liquids. The SD900 Portable Sampler collects liquid samples at specified intervals and keeps the samples in bottles or containers. Use the sampler for a wide variety of aqueous applications and for toxic pollutants and suspended solids. Set up the sampler with different retainers, bottles or containers. Refer to Figure 1. Figure 1 Product overview 1 Compact base 6 Controller connections 2 Standard insulated base 7 Pump 3 Center section 8 Controller 4 Power source 9 Liquid sensor 5 Top cover 10 Standard insulated base for 21-L (5.5 gal) container Product components Make sure that all components have been received. Refer to Figure 2. If any items are missing or damaged, contact the manufacturer or a sales representative immediately. English 7 Figure 2 Product components 1 Base (Standard, compact or composite) 6 Battery charger (optional) 2 Components for a single-bottle option (bottle and support can change) 7 AC power supply (optional) 3 Center section with controller 8 Strainer 4 Battery (optional) 9 Intake tubing, vinyl or Teflon-lined 5 Top cover 10 Components for a multiple-bottle option (bottles and retainers can change) Installation DANGER Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document. Confined space precautions DANGER Explosion hazard. Training in pre-entry testing, ventilation, entry procedures, evacuation/rescue procedures and safety work practices is necessary before entering confined spaces. The information that follows is supplied to help users understand the dangers and risks that are associated with entry into confined spaces. On April 15, 1993, OSHA's final ruling on CFR 1910.146, Permit Required Confined Spaces, became law. This standard directly affects more than 250,000 industrial sites in the United States and was created to protect the health and safety of workers in confined spaces. Definition of a confined space: A confined space is any location or enclosure that has (or has the immediate potential for) one or more of the following conditions: • An atmosphere with an oxygen concentration that is less than 19.5% or more than 23.5% and/or a hydrogen sulfide (H2S) concentration that is more than 10 ppm. • An atmosphere that can be flammable or explosive due to gases, vapors, mists, dusts or fibers. • Toxic materials which upon contact or inhalation can cause injury, impairment of health or death. 8 English Confined spaces are not designed for human occupancy. Confined spaces have a restricted entry and contain known or potential hazards. Examples of confined spaces include manholes, stacks, pipes, vats, switch vaults and other similar locations. Standard safety procedures must always be obeyed before entry into confined spaces and/or locations where hazardous gases, vapors, mists, dusts or fibers can be present. Before entry into a confined space, find and read all procedures that are related to confined space entry. Mechanical installation Site installation guidelines DANGER Explosion hazard. The instrument is not approved for installation in hazardous locations. Refer to the guidelines that follow for the site location evaluation. • Obey all the safety precautions if the sampler is installed in a confined space. Refer to Confined space precautions on page 8. • Make sure that the temperature at the location is in the specification range. Refer to Specifications on page 3. • Install the sampler on a level surface or hang the sampler with the suspension harness, the support bracket or the spanner bar. Refer to Install the sampler in a manhole on page 9 and to the applicable installation documentation. • As near the sample source as possible to decrease analysis delay. Refer to Plumb the sampler on page 12. • For limitations on transport velocity and maximum vertical lift, refer to Specifications on page 3. Install the sampler in a manhole Install the sampler above the sample water in a manhole. Install the sampler with a spanner bar or a support bracket. Install the spanner bar inside the manhole. The spanner bar is supported by pressure against the walls. The support bracket has the same width as the manhole cover. Install the support bracket directly below the cover for support.Refer to the expanded manual on the manufacturer's website.Refer to the documentation supplied with the accessories to install the sampler. Install the distributor or full-bottle shutoff (optional) The distributor or full-bottle shutoff assembly is typically installed at the factory. Refer to the distributor or full-bottle shutoff documentation for installation. Prepare the sampler Clean the sample bottles Clean the sample bottles and caps with a brush, water and a mild detergent. Flush the containers with fresh water followed by a distilled water rinse. Install a single bottle Use a single bottle to collect one composite sample. When the bottle is full, the full bottle shut-off stops the sample program. 1. Clean the sample bottles. Refer to Clean the sample bottles on page 9. 2. Install the sample bottle as shown in Figure 3. English 9 Figure 3 Single bottle installation 1 Polyethylene bottle, 10 L (2.5 gal) (1918) 6 Polyethylene bottle, 21 L (5.5 gal) (6494) 2 Glass bottle, 10 L (2.5 gal) (6559) 7 Compact base (8975) 3 Support (1502) 8 Standard insulated base (8976) 4 Polyethylene bottle, 15 L (4 gal) (1367) 9 Standard insulated base for 21 L (5.5 gal) bottle (8561) 5 Polyethylene bottle, 19 L (5 gal) (6498) Install the first bottle for multiple sample collections Use multiple bottles to collect samples into separate bottles or into more than one bottle. The distributor moves the sample tube above each bottle. Install the bottles in the sampler base as shown in Figure 4. Install the first sample bottle (number 1) below the label in the sampler base. Install the remaining bottles in increasing numbers in the direction shown by the label. Refer to Figure 5 on page 12 for a diagram of necessary components. 10 English Figure 4 Bottle number 1 installation 1 Bottle number 1 location (compact base) 6 Elastic straps 2 Bottle number 1 location for 24 bottles 7 Standard insulated base for 21 L (5.5 gal) bottle 3 Bottle number 1 location for 8 bottles 8 Standad insulated base 4 Bottle number 1 location for 2 or 4 bottles 9 Compact base 5 Retainer Install multiple bottles When multiple bottles are installed, a distributor arm moves the sample tube over each bottle. Sample collection automatically stops when the specified number of samples have been collected. 1. Clean the sample bottles. Refer to Clean the sample bottles on page 9. 2. Install the first sample bottle (number 1) below the label in the sampler base. Refer to Install the first bottle for multiple sample collections on page 10. 3. Assemble the sample bottles as shown in Figure 5. For eight or more bottles, make sure that the first bottle goes next to the bottle one indicator in the clockwise direction. 4. Put the bottle assembly in the sampler. For eight or more bottles, align the wires in the slots in the bottom tray. English 11 Figure 5 Multiple bottle installation 1 Positioner/retainer (2347) 8 Glass bottles, 3.8 L (1 gal), (4x) (2216) or polyethylene bottles, 3.8 L (1 gal), (4x) (2217) 2 Retainer (1422) 9 Glass bottles, 1.9 L (0.5 gal), (8x) (1118) or polyethylene bottles, 2.3 L (0.61 gal), (8x) (657) 3 Retainer (2190) 10 Polyethylene bottles, 1 L (0.26 gal), (24x) (737) 4 Retainer (2189) 11 Glass bottles, 350 mL (11.83 oz), (24x) (732) 5 Glass bottles, 950 mL (32.12 oz) (8x) (2348) 12 Compact base (8975) 6 Polyethylene bottles, 575 mL (19.44 oz), (24x) (1369) 13 Standard base (8976) 7 Glass bottles, 3.8 L (1 gal), (2x) (2214) or polyethylene bottles, 3.8 L (1 gal), (2x) (2215) Plumb the sampler Install the strainer in the middle of the sample stream (not near the surface or the bottom) to make sure that a representative sample is collected. Refer to Figure 6. Refer to Figure 7 for the intake tube installation. 1. Connect the tubing to the sampler as shown in Figure 7. Note: Use connection kit 2186 if Teflon-lined tubing is used. 2. Install the intake tube and strainer in the main stream of the sample source where the water is turbulent and well-mixed. • Make the intake tube as short as possible. • Keep the intake tube at a maximum vertical slope so that the tube drains completely between samples. Note: If a vertical slope is not possible or if the tube is pressurized, disable the liquid sensor. Calibrate the sample volume manually. Refer to Calibrate the sample volume manually on page 22. • Make sure that the intake tube is not pinched. 12 English Figure 6 Instrument installation 1 Strainer 3 Intake tube 2 Vertical lift Figure 7 Intake tube installation Electrical installation Controller connections WARNING Electrical shock hazard. Externally connected equipment must have an applicable country safety standard assessment. Figure 8 shows the connections that can be made to the controller. English 13 Figure 8 Controller connectors 1 Auxiliary device 4 SDI-12 device option 2 Power supply 5 Distributor/full bottle shut-off 3 Serial communications Connect a flow meter Connect a flow meter to the controller to start or stop the sampler when the sample flow goes above or below a specified value. Items to collect: • Multi-purpose full cable for Sigma flow meters (or 980 half cable for the model 980 flow meter). • Optional splitter for additional connections. Two or more splitters can be connected in series. 1. Connect one end of the cable to the flow meter. For the model 980 flow meter, refer to the model 980 flow meter user manual. 2. Connect the other end of the cable to the auxiliary device connector on the controller. Note: If the flow meter has a 6-pin cable, use the 6-pin to 7-pin adapter cable. Connect a PC or communications network Connect a PC or a communications network to the controller to transfer data or to configure the sampler. Items to collect: • Serial cable, 7-pin RS232 to DB-9 (refer to Figure 9). Connections: B to 5 (signal ground); D to 3 (RCD); F to 2 (TXD); G (ground). 1. Connect one end of the serial cable to the communications device or network. 2. Connect the other end of the cable to the serial communications connector on the controller. Figure 9 Serial communications cable 14 English Connect the sampler to power DANGER Electrocution hazard. If this equipment is used outdoors or in potentially wet locations, a Ground Fault Circuit Interrupt (GFCI/GFI) device must be used for connecting the equipment to its main power source. DANGER Fire hazard. Install a 15 A circuit breaker in the power line. A circuit breaker can be the local power disconnect, if located in close proximity to the equipment. DANGER Electrocution hazard. Protective Earth Ground (PE) connection is required. WARNING Electrocution hazard. Make sure that there is easy access to the local power disconnect. WARNING Electric shock hazard. The power supply can overheat when the time between pump cycles is too short. NOTICE Make sure that the pump cycle time will not cause the power supply to overheat. Refer to Calculate the pump cycle time on page 16. Connect the sampler to a battery or to an AC power supply. Refer to the illustrated steps that follow. English 15 Calculate the pump cycle time The pump must stay off for a minimum time period between sample cycles so that the power supply does not overheat. 1. Calculate the total amount of time that the pump will run continuously during a sample cycle. Include all stages: pre-purge, intake rinse, sample, sample retries and post-purge. 2. Find the minimum amount of time that the pump must stay off between sample cycles from Figure 10. 3. Make sure that the pacing or time interval in the sampler program lets the pump stay off between sample cycles for the minimum amount of time. Figure 10 Pump on and off times at 50 °C (122 ºF) 1 Time (in minutes) that the pump must stay off 2 Time (in minutes) that the pump is on continuously 3 Example: if the pump is on continuously for 5 minutes, the pump must stay off for 15 minutes. Set the power to on Push the power button to set the power to on or off. Make sure that the power supply (AC power or battery power) is correctly installed. User interface and navigation The controller keypad is shown in Figure 11. Use the MENU key to set up sampling programs, configure the controller settings or complete the diagnostic tests. Use the arrows, ENTER and BACK keys to scroll through the menu, make selections and enter values. Look for arrows on the bottom or side of the display screen to know when more options are available. 16 English Figure 11 Controller keypad 1 POWER 5 MENU 9 STATUS 2 VOLUME CALIBRATION 6 ENTER 10 RUN/HALT PROGRAM 3 STOP 7 BACK 11 MANUAL OPERATION 4 LED 8 ARROW KEYS Operation Configure the system settings 1. Push MENU. 2. Go to SYSTEM SETUP. 3. Change the applicable options. Option Description Time/date setup Set the time (24-hour format) and date. Communication Select the baud rate (19200, 38400, 57600 or 115200) and protocol (Modbus RTU or ASCII) for the serial port. Setup base Select the sampler base (All weather refrigerated, refrigerated, portable standard or portable compact). Language Select the language of the controller. Liq sensor cal Calibrate the liquid sensor or set to default. Set contrast Adjust the contrast of the LCD screen. Pump duty cycle Change the pump duty cycle. Range: 50% to 100% (default: 100%). Password setup Set up a password to control access to the program setup and the system setup menus. Select ENABLE>YES to set a new password or ENABLE>NO to activate the current password. If the password is set for the first time, use 900900 as the current password. English 17 Option Description Tubing life Set a reminder when the pump tubing is to be changed. Go to ENABLE>RESET CYCLES to reset the pump cycle count to 0. Go to ENABLE>CYCLE LIMIT to change the number of pump cycles that is used for the reminder. Go to ENABLE>STATUS to show the current number of pump cycle counts. SDI–12 (optional) Configure the operation of an SDI–12 device. This option is shown only when a device is found. Set up a sampling program A sampling program includes all of the parameters that are necessary to collect samples automatically. Collect the samples at regular intervals or when the sampling program is complete. 1. Push MENU. 2. Go to PROGRAM SETUP>MODIFY ALL. The first screen is shown. 3. Select or enter the values for each parameter. Use the left and right arrow keys to move left or right. Use the up and down arrow keys to change a value. Change the units if necessary (gal or mL) with the arrow keys. Option Description Bottles Bottle quantity: the number of bottles in the sampler (1, 2, 4, 8 or 24). Full BTL shutoff: Enabled/Disabled. If enabled, the controller checks for bottle overflow conditions. Bottle volume: the volume capacity of each bottle. Range: 50–65000 mL (0.01 17.17 gal) Intake tubing Tubing length: the length of the intake tubing from the strainer to the liquid sensor. Range: 100–3000 cm or 3–99 ft. An accurate length is necessary to get an accurate sample volume. Intake tube type: the diameter and material of the intake tubing (¼ in. vinyl, 3/8 in. vinyl or 3/8 in. Teflon). Program delay Sample pacing/collection Enable/disable: when enabled, the sampling program starts at a specified time or number of counts. Date and time: the date and time when the program starts (24hour format). Counts: the number of counts or pulses from a flow meter when the program will start (1 to 9999 counts). If the sample pacing is later set to time, the program delay will be disabled. Time based—Refer to Time based collection on page 19. Note: When CTVV is used: • • • • • The liquid sensor is always enabled. The sample distribution is composite. The run mode is non-continuous. The first sample is taken after the interval. Flow source is always 4–20 mA input. Flow based—Refer to Flow based collection on page 20. Sample distribution Deliver samples to all>YES Each sample is delivered to all bottles. End after last: program stops after the last sample (1–999). Continuous: program continues until it is stopped manually. Deliver samples to all>NO Samples are delivered to a subset of bottles . Samples/bottle: the number of samples to be collected in each bottle (1–999). Bottles/sample: The number of bottles that will contain the same sample. Liquid sensor 18 English Enable or disabled. If disabled, the sample volume must be calibrated by time. Option Description Sample volume The volume of each sample in mL (1 gal = 3785.4 mL). If bottles/sample mode is used, each bottle receives a full sample volume (10–10,000 mL). Make sure that the sample volume does not exceed the bottle volume. Sample volumes are rounded to the nearest 10 mL. Intake rinses The number of intake tube rinses before a sample is collected (0–3). Sample retries The number of sampling tries after a sampling failure occurs (0–3). Site ID The name for the sampling location (up to 12 characters). The site ID is used as the preset name if the program is saved. Advanced sampling Goes to the advanced sampling menu. If necessary, configure the options in the Advanced Sampling menu. Run program The creation of the basic sampling program finishes. The sampler prompts the user to start or cancel the program. Time based collection 1. Select or enter the values for each option. Option Description Pacing interval Collect samples at regular time intervals, in hours and minutes (0:01–999:00). Take first sample The program starts immediately or after the first interval. Variable volume Specify whether the sample volumes vary (flow based). If yes, refer to Use variable volume on page 19. If no, refer to Do not use variable volume on page 20. Use variable volume Sample pacing is constant time variable volume (CTVV). 1. Select or enter the values for each option. Option Description Select unit Select the flow units to use. Map 4 mA input Enter the flow rate that agrees with 4 mA input. Map 20 mA input Enter the flow rate that agrees with 20 mA input. Average flow rate Specify the average flow rate for the site (1–999,999). The units are specified by the units selected above. Time volume desired Specify the total volume to be collected during the sample program (10–10,000 ml). English 19 Option Description Collection period Specify the total collection period in hours and minutes (0:01–999:00). Minimum sample volume If the flow volume is not sufficient to collect this minimum volume, the sample is skipped (10–10,000 ml). Do not use variable volume 1. Select or enter the values for the option. Option Description Take first sample Select whether the program starts immediately or after the first time interval has passed. Flow based collection Specify the flow source for flow based sampling: • Counts—Refer to Use counts for the flow source on page 20. • 4–20 mA input—Refer to Use 4–20 mA input for the flow source on page 20. Use counts for the flow source 1. Select or enter the values for each option. Option Description Take sample every Collect samples at regular flow intervals, in counts (1–9999 counts). Overrride time Collect a sample if the flow volume is very low. Time The maximum time between samples (0:01–999:00). The timer starts again after each sample is collected. Take first sample The program starts immediately or after the first time interval. Use 4–20 mA input for the flow source 1. Select or enter the values for each option. Option Description Select unit Selects the flow units to use. Map 4 mA input Enter the flow rate that would agree with the 4 mA input. Map 20 mA input Enter the flow rate that would agree with the 20 mA input. Take sample every Enter the total flow after which the sample has to be drawn. Override time Select enable to push a sample to be collected if the flow volume is unusually low. Time The maximum time between samples (0:01–999:00). The timer starts again after each sample is collected. Sample volume Enter the volume to be collected per sample. If bottles/sample mode is selected, each bottle receives a full sample volume (100–10,000 mL). This option is skipped if variable volume is selected. Restore the default settings The restore option sets all of the program parameters to the default settings. 20 English 1. Push MENU. 2. Go to PROGRAM SETUP>RESTORE. 3. Select YES. Start or stop a program Use the RUN/HALT PROGRAM key to start or stop a program. Up to 12 user-defined start/stop times/dates can be set. A program must be stopped before the program or system setup can be changed. 1. Push RUN/HALT PROGRAM. 2. Select one of the options. Option Description Start Starts the currently loaded program. The status changes to RUNNING. Halt Stops the program temporarily. The status changes to HALTED. Resume The program resumes from the point at which it was halted. The status changes to RUNNING. Start over The program starts from the beginning. The status changes to RUNNING. End program Stops the currently running program. The status changes to COMPLETE. Manual operation Use manual operation to collect a grab sample, move the distributor arm or operate the pump. For additional information, refer to the expanded version of this manual. Volume calibration The sample volume can be calibrated automatically with the liquid sensor or manually. When the liquid sensor is used for volume calibration, the sensor is adjusted to accurately measure all programmed volumes. If the liquid sensor is disabled, the volume for samples, rinses and stormwater must be calibrated manually. For typical applications, tap water can be used for the volume calibration. If the sample composition is significantly different from typical water samples, calibrate the sampler with the liquid to be collected. All programs must be stopped before calibration. Calibrate the liquid sensor The accuracy of the liquid sensor varies with the type of liquid that is sampled. For example, the volume of a highly turbid sample may not be as accurate as that of clear water. The liquid sensor can be calibrated with the sample water to improve volume accuracy. 1. Put the intake tubing into the sample water. 2. Push MENU. 3. Go to PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE to make sure that the liquid sensor is enabled. 4. Go to SYSTEM SETUP>LIQ SENSOR CAL. 5. Select PERFORM CAL. 6. Select START. The pump operates in reverse to purge the intake tube. The pump then operates in the forward direction. 7. Look at the sample flow in the intake tube to make sure that the sample goes through the liquid sensor. 8. When the sample goes through the liquid sensor and no bubbles are present, push STOP. The intake tube is automatically purged. The liquid sensor calibration is complete. 9. Select DONE to exit or REPEAT to do the calibration again. English 21 Restore the default calibration The volume calibration for the liquid sensor can be returned to the default factory settings. 1. Select PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE to make sure that the liquid sensor is enabled. 2. Push VOLUME CALIBRATION and select RESET CAL. The volume calibration is set to the default settings. The message “value has been reset to 0” will be shown. Calibrate the sample volume with the liquid sensor Calibrate the sample volume with the liquid sensor enabled to adjust sample volumes slightly. 1. Go to PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE to make sure that the liquid sensor is enabled. 2. Put the intake tubing into the sample water or tap water. Note: Use the sample water for best accuracy. 3. Put the tubing from the outlet side of the pump into a graduated cylinder. 4. Push VOLUME CALIBRATION and select CALIBRATION. 5. Select START. The pump operates in reverse to purge the intake tube. The pump then operates in the forward direction and collects the sample volume that is specified in the sampling program. The pump operates in reverse to purge the intake tube. 6. When the sample has been collected, select DONE. Compare the volume of the sample in the graduated cylinder with the sample volume that is entered in the program setup menu. 7. If the volume that was collected is different from the sample volume in the current program, enter the volume that was actually collected. Push ENTER. The sensor is adjusted to accurately measure all programmed volumes. 8. Select START to start the sampling program or CANCEL to exit the volume calibration menu. Calibrate the sample volume manually When the liquid sensor is disabled, all volumes must be calibrated manually by time. The sample volume is calibrated for the volume specified in the current program. If the sample volume is changed in the program, the sample volume must be calibrated again for the new volume. 1. Select PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>DISABLE to make sure that the liquid sensor is disabled. 2. Put the intake tubing into the sample water or tap water. 3. Put the tubing from the outlet side of the pump into a graduated cylinder. 4. Push VOLUME CALIBRATION. If more than one volume is shown, select one of the volumes for calibration. 5. Select START. The pump operates in reverse to purge the intake tube. The pump then operates in the forward direction and starts to collect the sample. 6. Stop the pump at the selected volume or rinse location: • Sample volume: push STOP when the volume that is specified in the main program is collected. • Rinse: push STOP when the liquid in the intake tubing gets to the liquid detector. • Storm volume: push STOP when the volume specified in the stormwater program is collected. 7. Select DONE to go to the next volume calibration or REPEAT to do the calibration again. 8. When all volumes have been calibrated, select DONE. 9. Connect the outlet tubing to the tube fitting on the sampler. 10. Select START to start the sampling program or CANCEL to exit. 22 English Verify the sample volume To make sure that the sample volume is accurate, take a grab sample. Do not go back into calibration to check the volume since the volume compensation is reset to zero at the start of a calibration. 1. 2. 3. 4. 5. 6. 7. 8. Push MANUAL OPERATION. Select GRAB SAMPLE. Put the intake tubing into the sample water. Put the tubing from the outlet side of the pump into a graduated cylinder. If the liquid sensor is enabled, enter the volume to be verified. If the liquid sensor is disabled, enter the volume that is specified in the program. Push ENTER. The pump cycle starts (purge-sample-purge). Compare the volume that was collected in the graduated cylinder to the expected volume. If the collected volume is not correct, do the volume calibration again. Maintenance DANGER Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document. DANGER Electrocution hazard. Remove power from the instrument before doing maintenance or service activities. WARNING Biohazard exposure. Obey safe handling protocols during contact with sample bottles and sampler components. WARNING Multiple hazards. The technician must make sure that the equipment operates safely and correctly after maintenance procedures. Clean the instrument CAUTION Fire hazard. Do not use flammable agents to clean the instrument. NOTICE Do not clean the controller compartment heater with liquids of any kind. If water is not sufficient to clean the controller and the pump, disconnect the controller and move the controller away from the sampler. Allow sufficient time for the controller and pump to dry before the parts are re-installed and put back into service. Clean the sampler as follows: • Sampler enclosure—Clean the internal and external surfaces with a damp cloth and a mild detergent. Do not use abrasive cleaners or solvents. English 23 • Sample bottles—Clean the bottles with a brush and water and a mild detergent. Rinse with fresh water and then with distilled water. Use an autoclave to clean glass bottles if necessary. Replace the desiccant A desiccant cartridge in the controller absorbs moisture and prevents corrosion. Monitor the desiccant color through the window (Figure 12). Fresh desiccant is yellow. When the color is green, replace the desiccant. 1. 2. 3. 4. 5. 6. Unscrew and remove the desiccant cartridge (Figure 12). Remove the plug and discard the spent desiccant. Fill the desiccant tube with fresh desiccant. Install the plug. Apply grease to the O-ring. Install the desiccant tube in the controller. Figure 12 Desiccant cartridge 1 Plug 3 O-ring 2 Desiccant tube 4 Desiccant window Pump maintenance CAUTION Pinch hazard. Remove power from the instrument before maintenance or service activities are done. Replace the pump tubing NOTICE Use of tubing other than that supplied by the manufacturer can cause excessive wear on mechanical parts and/or poor pump performance. Examine the pump tubing for wear where the rollers rub against the tubing. Replace the tubing when the reminder message is shown or when the tubing shows signs of wear. Pre-requisites: • Pump tubing—pre-cut or bulk 4.6 m or 15.2 m (15 ft or 50 ft) 1. Disconnect the power to the controller. 2. If the bulk tubing is used, cut the tubing and add alignment dots. Refer to the expanded manual on the manufacturer's website. 3. Remove the pump tubing as shown in the illustrated steps that follow. 4. Clean the silicone residue from the interior of the pump housing and from the rollers. 5. Install the new pump tubing as shown in the illustrated steps that follow. 24 English English 25 Clean the rotor Clean the rotor, pump tube tracks and pump housing with a mild detergent. Refer to Replace the pump tubing on page 24 and the illustrated steps that follow. Replace the distributor arm tube The distributor arm moves over each bottle during multiple bottle sampling. Replace the tube in the distributor arm when the tube is worn. Make sure that the correct tube is used for the correct distributor and distributor arm. Note: The distributor tubing is not the same as the pump tubing. The pump tubing installed in the distributor assembly can damage the distributor. Also, samples can be missed because the distributor arm cannot move easily. 1. Remove the tube from the distributor arm and from the ceiling of the center section. 2. Insert the new tube into the distributor arm. Make sure that the tube is aligned with the dimensions shown in Figure 13. 3. Insert the other end of the tube into the fitting on the ceiling of the center section. 4. Complete the diagnostic test for the distributor to make sure that the operation is correct. Figure 13 Distributor assembly 1 Nozzle 3 Distributor motor 2 Distributor arm 4 Shaft 26 English 5 Distributor tube Fuse replacement The fuse is a not a user-serviceable item. Contact technical support. A blown fuse can be an indication that the sampler has a problem and that service is necessary. English 27 Table des matières Caractéristiques à la page 28 Fonctionnement à la page 42 Généralités à la page 30 Maintenance à la page 49 Installation à la page 33 Informations supplémentaires Des informations supplémentaires sont disponibles sur le site Web du fabricant. Caractéristiques Les caractéristiques techniques peuvent être modifiées sans préavis. Echantillonneur portable SD900 Caractéristique Base standard Base compacte Base composite Dimensions Diamètre : 50,5 cm (19,9 po) Diamètre : 44,1 cm (17,4 po) Diamètre : 50,28 cm (19,8 po) Hauteur : 61 cm (24 po) Hauteur : 79,75 cm (31,4 po) 12,2 kg (27 lb) avec des bouteilles en polyéthylène de 575 ml (19,44 oz) (24x) 15 kg (36 lb) avec des bouteilles en verre de 950 ml (32,12 oz) (12x) Hauteur : 69,4 cm (27,3 po) Poids 15 kg (35,6 lb) avec des bouteilles en polyéthylène de 1 l (24x) 14,8 kg (32,6 lb) avec un récipient en polyéthylène de 10 l (2,5 gal) (1x) 12,9 kg (28,3 lb) avec un récipient en polyéthylène de 10 l (2,5 gal) (1x) Boîtier ABS résistant aux impacts, construction en 3 sections, base à double paroi avec isolation de 2,54 cm (1 po) ; contact direct entre la bouteille et la glace. Température de l'échantillon 0 à 60 °C (32 à 140 °F) Arrêtoirs Acier inoxydable 316 dans la taille standard, haute vitesse ou bas profil pour des applications peu profondes et acier inoxydable 316 ou en Téflon® dans la taille standard Tuyauterie d'admission de l'échantillon 9,5 mm (3/8 po) DI en vinyle ou polyéthylène doublé en Téflon Capacité de la bouteille d'échantillon Bouteilles de 1 l (0,26 gal) en polyéthylène et/ou de 350 ml (11,83 oz) en verre (24x) Bouteilles en verre de Bouteilles de 2,3 l (0,6 gal) en 950 ml (32,12 oz) (8x) polyéthylène et/ou de 1,9 l (0,5 gal) Bouteilles en polyéthylène en verre (8x) de 10 l (2,5 gal) (1x) Bouteilles de 3,8 l (1 gal) en Bouteilles en verre de 10 l polyéthylène et/ou de 3,8 l (1 gal) (2,5 gal) (1x) en verre (4x) Bouteilles de 3,8 l (1 gal) en polyéthylène et/ou de 3,8 l (1 gal) en verre (2x) Récipient de 21 l (5,5 gal) ou de 15 l (4 gal) en polyéthylène composite ou de 20 l (5,25 gal) en polyéthylène, ou de 10 l (2,5 gal) en polyéthylène, ou de 10 l (2,5 gal) en verre (1x) 28 Français Bouteilles en polyéthylène de 575 ml (19,44 oz) (24x) Bouteilles en polyéthylène de 21 l (5,5 gal) (1x) Contrôleur SD Caractéristique Détails Dimensions (L x H x P) 29,2 x 17,1 x 26,4 cm (11½ x 6¾ x 103/8 po.) Poids 4.2 kg (9.26 lb) Boîtier Alliage PC/ABS, NEMA 4X, 6, IP 67 Alimentation requise 15 V c.c. fournis par une alimentation 8754500 ; 15 V c.c. fournis par une alimentation complète Protection contre les surcharges 7 A, fusible de ligne c.c. pour la pompe Température de fonctionnement 0 à 50 °C Température de stockage –30 à 60 °C (–22 à 140 °F) Humidité de stockage/de fonctionnement 100% de condensation Pompe Grande vitesse péristaltique, avec rouleaux en Nylatron montés à ressort Corps de pompe IP37 Tuyau de pompe 9,5 mm DI x 15,9 mm DE (3/8 po. DI x 5/8 po. DE) en silicone Durée de vie de la pompe 20 000 cycles d'échantillonnage avec : 1 l (0,3 gal) de volume d'échantillon, 1 rinçage, 6 minutes d'intervalle, 4,9 m (16 pieds) du tube d'admission de 3/8 po., 4,6 m (15 pieds) de levée verticale, température d'échantillon de 21 °C (70 °F) Levée verticale de l'échantillon Maximum 8,5 m (28 pieds) pour : 8,8 m (29 pieds) du tube d'admission en vinyle de 3/8 po. au niveau de la mer à 20-25 °C (68-77 °C) Débit de pompe 4,8 l/min (1,25 gpm) avec une levée verticale de 1 m (3 ft) avec un tube d'admission de 3/8 po. typique Sample volume (Volume d’échantillon) Programmable par incréments de 10 ml (0,34 oz) de 10 à 10 000 ml (3,38 oz à 2,6 gal) Répétabilité du volume d'échantillon (typique) ±5 % de volume d'échantillon de 200 ml par détection de liquide non étalonné avec 4,6 m (15 pieds) de levée verticale, 4,9 m (16 pieds) du tube d'admission en vinyle de 3/8 po., bouteille seule, fermeture de bouteille pleine à température ambiante et avec une altitude de 1 524 m (5 000 pieds) Précision du volume d'échantillon (typique) ±10 % de volume d'échantillon de 200 ml par détection de liquide non étalonné avec : 4,6 m (15 pieds) de levée verticale, 4,9 m (16 pieds) du tube d'admission en vinyle de 3/8 po. bouteille seule, fermeture de bouteille pleine à température ambiante et avec une altitude de 1 524 m (5 000 pieds) Modes d'échantillonnage Rythme : durée fixe, débit fixe, durée variable, débit variable, volume variable à intervalle fixe (CTVV). Reportez-vous à la section sur l'échantillonnage de volume variable à intervalle fixe (CTVV) dans la version complète de ce manuel. Distribution : composite bouteille unique, composite multi-bouteilles, discrète multi-bouteilles, bouteilles par échantillon, échantillons par bouteille, combinaison de bouteilles par échantillon, échantillons par bouteille Modes d'exécution Continu ou non continu avec nombre d'échantillons entré par l'utilisateur Programmes multiples Enregistre jusqu'à trois programmes d'échantillonnage Vitesse du transfert (typique) 0,9 m/s (2,9 pieds/s) avec : 4,6 m (15 pieds) de levée verticale, 4,9 m (16 pieds) du tube d'admission en vinyle de 3/8 po., 21 °C (70 °F) et 1 524 m (5 000 pieds) d'altitude Capteur de liquide Ultrasonique. Corps : Ultem agréé NSF ANSI norme 51, conforme USP classe VI ® Français 29 Caractéristique Détails Historique d'échantillonnage Jusqu'à 510 enregistrements Purge d'air Air purgé automatiquement avant et après chaque échantillon. La durée compense automatiquement des variations des longueurs de tube d'admission. Journal événements 510 enregistrements Connexions Alimentation, auxiliaire, communication série, distributeur, SDL-12 Matériaux immergés Acier inoxydable, polyéthylène, Teflon, Ultem, silicone Communications RS232, Modbus, SDI-12 Garantie 1 an Généralités En aucun cas le constructeur ne saurait être responsable des dommages directs, indirects, spéciaux, accessoires ou consécutifs résultant d'un défaut ou d'une omission dans ce manuel. Le constructeur se réserve le droit d'apporter des modifications à ce manuel et aux produits décrits à tout moment, sans avertissement ni obligation. Les éditions révisées se trouvent sur le site Internet du fabricant. Consignes de sécurité AVIS Le fabricant décline toute responsabilité quant aux dégâts liés à une application ou un usage inappropriés de ce produit, y compris, sans toutefois s'y limiter, des dommages directs ou indirects, ainsi que des dommages consécutifs, et rejette toute responsabilité quant à ces dommages dans la mesure où la loi applicable le permet. L'utilisateur est seul responsable de la vérification des risques d'application critiques et de la mise en place de mécanismes de protection des processus en cas de défaillance de l'équipement. Veuillez lire l'ensemble du manuel avant le déballage, la configuration ou la mise en fonctionnement de cet appareil. Respectez toutes les déclarations de prudence et d'attention. Le non-respect de cette procédure peut conduire à des blessures graves de l'opérateur ou à des dégâts sur le matériel. Assurez-vous que la protection fournie avec cet appareil n'est pas défaillante. N'utilisez ni n'installez cet appareil d'une façon différente de celle décrite dans ce manuel. Interprétation des indications de risques DANGER Indique une situation de danger potentiel ou imminent qui, si elle n'est pas évitée, entraîne des blessures graves, voire mortelles. AVERTISSEMENT Indique une situation de danger potentiel ou imminent qui, si elle n'est pas évitée, peut entraîner des blessures graves, voire mortelles. ATTENTION Indique une situation de danger potentiel qui peut entraîner des blessures mineures ou légères. AVIS Indique une situation qui, si elle n'est pas évitée, peut occasionner l'endommagement du matériel. Informations nécessitant une attention particulière. Etiquettes de mise en garde Lisez toutes les informations et toutes les étiquettes apposées sur l’appareil. Des personnes peuvent se blesser et le matériel peut être endommagé si ces instructions ne sont pas respectées. Un 30 Français symbole sur l'appareil est référencé dans le manuel et accompagné d'une déclaration de mise en garde. Ceci est le symbole d'alerte de sécurité. Se conformer à tous les messages de sécurité qui suivent ce symbole afin d'éviter tout risque de blessure. S'ils sont apposés sur l'appareil, se référer au manuel d'utilisation pour connaître le fonctionnement ou les informations de sécurité. Ce symbole indique un danger de pincement potentiel. Le matériel électrique portant ce symbole ne doit pas être mis au rebut dans les réseaux domestiques ou publics européens. Retournez le matériel usé ou en fin de vie au fabricant pour une mise au rebut sans frais pour l'utilisateur. Certification Règlement canadien sur les équipements causant des interférences radio, IECS-003, Classe A: Les données d'essai correspondantes sont conservées chez le constructeur. Cet appareil numérique de classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada. Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences. FCC part 15, limites de classe A : Les données d'essai correspondantes sont conservées chez le constructeur. L'appareil est conforme à la partie 15 de la règlementation FCC. Le fonctionnement est soumis aux conditions suivantes : 1. Cet équipement ne peut pas causer d'interférence nuisible. 2. Cet équipement doit accepter toutes les interférences reçues, y compris celles qui pourraient entraîner un fonctionnement inattendu. Les modifications de cet équipement qui n’ont pas été expressément approuvées par le responsable de la conformité aux limites pourraient annuler l’autorité dont l’utilisateur dispose pour utiliser cet équipement. Cet équipement a été testé et déclaré conforme aux limites définies pour les appareils numériques de classe A, conformément à la section 15 de la réglementation FCC. Ces limites ont pour but de fournir une protection raisonnable contre les interférences néfastes lorsque l’équipement fonctionne dans un environnement commercial. Cet équipement génère, utilise et peut irradier l'énergie des fréquences radio et, s'il n'est pas installé ou utilisé conformément au mode d'emploi, il peut entraîner des interférences dangereuses pour les communications radio. Le fonctionnement de cet équipement dans une zone résidentielle risque de causer des interférences nuisibles, dans ce cas l'utilisateur doit corriger les interférences à ses frais Les techniques ci-dessous peuvent permettre de réduire les problèmes d'interférences : 1. Débrancher l'équipement de la prise de courant pour vérifier s'il est ou non la source des perturbations 2. Si l'équipement est branché sur le même circuit de prises que l'appareil qui subit des interférences, branchez l'équipement sur un circuit différent. 3. Éloigner l'équipement du dispositif qui reçoit l'interférence. 4. Repositionner l’antenne de réception du périphérique qui reçoit les interférences. 5. Essayer plusieurs des techniques ci-dessus à la fois. Français 31 Présentation du produit DANGER Dangers chimiques ou biologiques. Si cet instrument est utilisé pour la surveillance d'un procédé de traitement et/ou d'un système de dosage de réactifs chimiques auxquels s'appliquent des limites réglementaires et des normes de surveillance motivées par des préoccupations de santé et de sécurité publiques ou de fabrication et de transformation d'aliments ou de boissons, il est de la responsabilité de l'utilisateur de cet instrument qu'il connaisse et applique les normes en vigueur et qu'il ait à sa disposition suffisamment de mécanismes pour s'assurer du bon respect de ces normes dans l'éventualité d'un dysfonctionnement de l'appareil. ATTENTION Risque d’incendie. Ce produit n'est pas adapté à l'utilisation avec des liquides inflammables. L'échantillonneur portable SD900 collecte des échantillons liquides à intervalles réguliers et conserve les échantillons dans des bouteilles ou des récipients. Utilisez l'échantillonneur pour un large éventail d'applications aqueuses, pour des substances toxiques et des solides en suspension. Configurez l'échantillonneur avec différentes bouteilles ou différents arrêtoirs ou récipients. Voir la Figure 1. Figure 1 Présentation du produit 1 Base compacte 6 Connexions du transmetteur 2 Base standard isolée 7 Pompe 3 Pièce centrale 8 Contrôleur 4 Source d'alimentation 9 Capteur de liquide 5 Couvercle supérieure 10 Base standard isolée pour un récipient de 21 l (5,5 gal) Composants du produit Assurez-vous d'avoir bien reçu tous les composants. Voir la Figure 2. Si des éléments manquent ou sont endommagés, contactez immédiatement le fabricant ou un représentant commercial. 32 Français Figure 2 Composants du produit 1 Base (standard, compacte ou composite) 6 Chargeur de batterie (en option) 2 Composants pour une option à bouteille unique (la bouteille et le support peuvent changer) 7 Alimentation secteur (en option) 3 Pièce centrale avec transmetteur 8 Crépine 4 Batterie (en option) 9 Tube d'admission, en vinyle ou doublé en téflon 5 Couvercle supérieure 10 Composants pour une option à plusieurs bouteilles (les bouteilles et les arrêtoirs peuvent changer) Installation DANGER Dangers multiples. Seul le personnel qualifié doit effectuer les tâches détaillées dans cette section du document. Précautions concernant l'espace confiné DANGER Risque d’explosion Une formation portant sur les tests de pré-entrée, la ventilation, les procédures d'entrée, les procédures d'évacuation/de sauvetage et les mesures de sécurité est nécessaire avant d'entrer dans des lieux confinés. Les informations suivantes sont fournies dans le but d'aider les utilisateurs à appréhender les dangers et les risques associés aux espaces confinés. Le 15 avril 1993, le règlement final de l'OSHA concernant le CFR 1910.146, Permit Required Confined Spaces (Espaces confinés nécessitant l'autorisation), est devenue une loi. Cette norme affecte directement plus de 250 000 sites industriels aux Etats-Unis et a été rédigée dans le but de protéger la santé et la sécurité des travailleurs en espace confiné. Définition d'un espace confiné : Tout endroit ou clôture qui présente (ou est susceptible de présenter) une ou plusieurs des conditions suivantes : • Une atmosphère qui contient une concentration d'oxygène inférieure à 19,5 % ou supérieure à 23,5 % et/ou une concentration de sulfure d'hydrogène (H2S) supérieure à 10 ppm. • Une atmosphère qui peut être inflammable ou explosive en présence de gaz, vapeurs, brumes, poussières ou fibres. Français 33 • Des matériaux toxiques qui, en cas de contact ou d'inhalation, sont susceptibles d'occasionner des blessures, des problèmes de santé ou la mort. Les espaces confinés ne sont pas conçus pour l'occupation humaine. Les espaces confinés disposent d’un accès limité et présentent des risques connus ou potentiels. Les trous d’homme, les colonnes, les tuyaux, les cuves, les chambres de commutation et autres emplacements similaires sont des exemples d’espaces confinés. Il convient de toujours suivre les procédures de sécurité standard avant d'entrer dans des espaces et/ou des endroits confinés soumis à des gaz dangereux, des vapeurs, des brumes, des poussières ou des fibres Avant de pénétrer dans un espace confiné, veuillez lire l'ensemble des procédures liées à l'accès. Installation mécanique Conseils d'installation sur site DANGER Risque d’explosion L'instrument n'est pas homologué pour une installation dans des zones dangereuses. Reportez-vous aux conseils ci-dessous pour l'évaluation de l'emplacement du site. • Suivez toutes les précautions de sécurité si l'échantillonneur est installé dans un espace confiné. Voir la Précautions concernant l'espace confiné à la page 33. • Vérifiez que la température de l'emplacement ne dépasse pas la plage spécifiée. Voir la Caractéristiques à la page 28. • Placez l'échantillonneur sur une surface plane ou suspendez l'échantillonneur à l'aide des sangles de suspension, du palier de support ou de la clef tricoise. Voir la Installez l'échantillonneur dans le trou d'homme à la page 34 et la documentation applicable pour l'installation. • Aussi près que possible de la source d'échantillon pour réduire le décalage d'analyse. Voir la Branchement de l'échantillonneur à la page 37. • Pour connaître les limitations de vitesse durant le transport et la levée verticale maximale, voir la Caractéristiques à la page 28. Installez l'échantillonneur dans le trou d'homme Installezl'échantillonneurau-dessus de l'eau de l'échantillon dans le trou d'homme. Installezl'échantillonneuravec une clef tricoise ou un palier de support. Installez la clef tricoise dans le trou d'homme. La clef tricoise est maintenue par pression contre le mur. Le palier de support comporte la même largeur que le couvercle de trou d'homme. Installez le palier de support directement sous le couvercle pour le maintenir.Consultez le manuel complet sur le site Web du fabricant.Consultez la documentation fournie avec les accessoires pour installer l'échantillonneur. Installez le distributeur ou la fermeture de bouteille pleine (en option) L'ensemble du distributeur ou de la fermeture de bouteille pleine est généralement monté en usine. Consultez la documentation du distributeur ou de la fermeture de bouteille pleine pour l'installation. Préparation de l'échantillonneur Nettoyez les bouteilles de l'échantillon Nettoyez les capuchons et bouteilles d'échantillon avec une brosse, de l'eau et un détergent doux. Rincez les conteneurs à l'eau douce, puis à l'eau distillée. Installation d'une bouteille Utilisez une seule bouteille pour prélever un échantillon composite. Lorsque la bouteille est pleine, le système de fermeture de bouteille pleine interrompt le programme d'échantillonnage. 1. Nettoyez les bouteilles de l'échantillon. Voir Nettoyez les bouteilles de l'échantillon à la page 34. 2. Installez la bouteille d'échantillon tel qu'illustré sur la Figure 3. 34 Français Figure 3 Installation de bouteille unique 1 Bouteille en polyéthylène, 10 l (2,5 gal) (1918) 6 Bouteille en polyéthylène, 21 l (5,5 gal) (6494) 2 Bouteille en verre, 10 l (2,5 gal) (6559) 7 Base compacte (8975) 3 Service après-vente (1502) 8 Base standard isolée (8976) 4 Bouteille en polyéthylène, 15 l (4 gal) (1367) 9 Base standard isolée pour une bouteille de 21 l (5,5 gal) (8561) 5 Bouteille en polyéthylène, 19 l (5 gal) (6498) Installez la première bouteille pour collecter des échantillons Utilisez des bouteilles multiples pour collecter des échantillons dans des bouteilles séparées ou dans plusieurs bouteilles. Le distributeur déplace le tube de prélèvement au-dessus de chaque bouteille. Positionnez les bouteilles dans l'embase de l'échantillonneur comme indiqué sur la Figure 4. Installez la première bouteille de l'échantillon (numéro 1) sous l'étiquette dans l'embase de l'échantillonneur. Installez les bouteilles restantes en nombre croissant dans le sens indiqué par l'étiquette. Voir sur la Figure 5 à la page 37 le schéma des composants requis. Français 35 Figure 4 Installation de la bouteille numéro 1 1 Emplacement de la bouteille numéro 1 (base compacte) 6 Sangles élastiques 2 Emplacement de la bouteille numéro 1 pour 24 bouteilles 7 Base standard isolée pour une bouteille de 21 l (5,5 gal) 3 Emplacement de la bouteille numéro 1 pour 8 bouteilles 8 Base standard isolée 4 Emplacement de la bouteille numéro 1 pour 2 ou 4 bouteilles 9 Base compacte 5 Arrêtoir Installation de plusieurs bouteilles Lorsque plusieurs bouteilles sont installées, un bras de distribution déplace le tube d'échantillon audessus de chaque bouteille. Le prélèvement d'échantillons s'interrompt automatiquement lorsque le nombre spécifié d'échantillons a été prélevé. 1. Nettoyez les bouteilles de l'échantillon. Voir Nettoyez les bouteilles de l'échantillon à la page 34. 2. Installez la première bouteille de l'échantillon (numéro 1) sous l'étiquette dans l'embase de l'échantillonneur. Voir Installez la première bouteille pour collecter des échantillons à la page 35. 3. Assemblez les bouteilles d'échantillon tel qu'illustré sur la Figure 5. Pour huit bouteilles ou plus, assurez-vous que la première bouteille se place à côté de l'indicateur de la première bouteille dans le sens des aiguilles d'une montre. 4. Placez l'assemblage de bouteilles dans l'échantillonneur. Pour huit bouteilles ou plus, alignez les câbles dans les emplacements du plateau inférieur. 36 Français Figure 5 Installation de bouteilles multiples 1 Positionneur/arrêtoir (2347) 8 Bouteilles en verre, 3,8 l (1 gal), (4x) (2216) ou bouteilles en polyéthylène, 3,8 l (1 gal), (4x) (2217) 2 Arrêtoir (1422) 9 Bouteilles en verre, 1,9 l (0,5 gal), (8x) (1118) ou bouteilles en polyéthylène, 2,3 l (0,61 gal), (8x) (657) 3 Arrêtoir (2190) 10 Bouteilles en polyéthylène, 1 l (0,26 gal), (24x) (737) 4 Arrêtoir (2189) 11 Bouteilles en verre, 350 ml (11,83 oz) (24x) (732) 5 Bouteilles en verre, 950 ml (32,12 oz) (8x) (2348) 12 Base compacte (8975) 6 Bouteilles en polyéthylène, 575 ml (19,44 oz), (24x) (1369) 13 Base standard (8976) 7 Bouteilles en verre, 3,8 l (1 gal), (2x) (2214) ou bouteilles en polyéthylène, 3,8 l (1 gal), (2x) (2215) Branchement de l'échantillonneur Monter la crépine au milieu du jet de l'échantillonneur (pas près de la surface ou près du fond) pour s'assurer qu'un échantillon représentatif est collecté. Voir la Figure 6. Voir la Figure 7pour installer le tube d'admission. 1. Connectez le tube à l'échantillonneur tel qu'illustré sur la Figure 7. Remarque : Utilisez le kit de connexion 2186 si un tube en Téflon est utilisé. 2. Mettez le tube d'admission et la crépine dans le flux principal de la source de l'échantillon où l'eau est turbulente et bien mélangée. • Raccourcissez autant que possible le tube d'admission. • Placez le tube d'admission verticalement afin que le tube se vide entièrement entre les échantillons. Remarque : si une pente verticale est impossible ou si le tube est pressurisé, désactivez le capteur de liquide. Etalonnage manuel du volume de l'échantillon. Voir la Etalonnage manuel du volume de l'échantillon à la page 48. Français 37 • Assurez-vous que le tube d'admission n'est pas pincé. Figure 6 Installation de l'instrument 1 Crépine 3 Tube d'admission 2 Relevage vertical Figure 7 Installation du tube d'admission Installation électrique Connexions du transmetteur AVERTISSEMENT Risque d'électrocution. Tout équipement externe relié doit avoir fait l'objet d'un contrôle de sécurité conformément aux normes nationales applicables. La Figure 8 illustre les connexions pouvant être effectuées avec le transmetteur. 38 Français Figure 8 Connecteurs du transmetteur 1 Appareil auxiliaire 4 Appareil SDI-12 en option 2 Alimentation 5 Distributeur/fermeture de bouteille pleine 3 Communications série Branchement d'un débitmètre Branchez un débitmètre au transmetteur pour démarrer ou interrompre l'échantillonneur lorsque le flux d'échantillon est supérieur ou inférieur à la valeur spécifiée. Articles à réunir : • Câble universel complet pour les débitmètres Sigma (ou demi-câble 980 pour le débitmètre de modèle 980). • Répartiteur en option pour les branchements additionnels. Deux répartiteurs ou plus peuvent être connectés en série. 1. Connectez une extrémité du câble audébitmètre. Pour le modèle de débitmètre 980, reportezvous au manuel de l'utilisateur. 2. Branchez l'autre extrémité du câble au connecteur d'appareil auxiliaire du transmetteur. Remarque : si le débitmètre dispose d'une câble à 6 broches, utilisez le câble d'adaptateur 6 à 7 broches. Connexion d'un PC ou d'un réseau de communication Connectez un PC ou un réseau de communication au transmetteur pour transférer des données ou configurer l'échantillonneur. Articles à réunir : • Câble série, RS232 vers DB-9 7 broches (voir Figure 9). Connexions : B vers 5 (masse du signal) ; D vers 3 (RDC) ; F vers 2 (TXD) ; G (masse). 1. Reliez une extrémité du câble série au réseau ou à l'appareil de communication. 2. Branchez l'autre extrémité du câble au connecteur de communication série du transmetteur. Figure 9 Câble de communication série Français 39 Branchement de l'échantillonneur à l'alimentation DANGER Risque d'électrocution Si cet équipement est utilisé à l'extérieur ou dans des lieux potentiellement humides, un disjoncteur de fuite à la terre (GFCI/GFI) doit être utilisé pour le branchement de l'équipement à sa source d'alimentation secteur. DANGER Risque d’incendie. Installez un disjoncteur 15 A sur la ligne électrique. Le disjoncteur peut servir à couper l'alimentation localement s'il est situé tout près de l'équipement. DANGER Risque d'électrocution Un raccordement à la terre est nécessaire. AVERTISSEMENT Risque d'électrocution Assurez-vous de disposer d'un accès facile à la coupure d'alimentation locale. AVERTISSEMENT Risque de décharge électrique. L'alimentation peut surchauffer quand le délai entre les cycles de pompe est trop court. AVIS Assurez-vous que la durée du cycle de la pompe ne provoquera pas la surchauffe de l'alimentation électrique. Voir la Calcul de la durée des cycles de la pompe à la page 41. Connectez l'échantillonneur à une batterie ou à une alimentation secteur. Reportez-vous aux illustrations suivantes. 40 Français Calcul de la durée des cycles de la pompe La pompe doit rester arrêtée pendant une durée minimum entre les cycles d'échantillonnage, afin que l'alimentation électrique ne surchauffe pas. 1. Déterminez la durée totale pendant laquelle la pompe fonctionnera de façon continue au cours d'un cycle d'échantillonnage. Inclure toutes les étapes : pré-purge, rinçage de l'admission, échantillonnage, échantillonnages répétés et post-purge. 2. Déterminez la durée minimale pendant laquelle la pompe doit être arrêtée entre les cycles d'échantillonnage à l'aide de la Figure 10. 3. Assurez-vous que le rythme ou l'intervalle du programme de l'échantillonneur permet l'arrêt de la pompe pendant la durée minimale déterminée entre les cycles d'échantillonnage. Figure 10 Durées d'arrêt/fonctionnement de la pompe à 50 °C (122 °F) 1 Durée (en minutes) pendant laquelle la pompe doit rester arrêtée 2 Durée (en minutes) pendant laquelle la pompe fonctionne de façon continue 3 Exemple : si la pompe fonctionne de façon continue pendant 5 minutes, la pompe doit rester arrêtée pendant 15 minutes. Mise sous tension Appuyez sur le bouton d'alimentation pour mettre l'appareil sous ou hors tension. Assurez-vous que l'alimentation (alimentation CA ou alimentation par batterie) est correctement installée. Interface utilisateur et navigation Le clavier du transmetteur est illustré à la Figure 11. Utilisez la touche MENU pour configurer les programmes d'échantillonnage et les paramètres du transmetteur ou réaliser les tests de diagnostic. Utilisez les flèches ainsi que les touches ENTER (Entrée) et BACK (Retour) pour parcourir le menu, sélectionner des éléments et saisir des valeurs. Vérifiez les flèches en bas ou à côté de l'écran pour savoir si d'autres options sont disponibles. Français 41 Figure 11 Clavier du transmetteur 1 MARCHE 5 MENU 9 STATUT 2 VOLUME CALIBRATION (Etalonnage du volume) 6 ENTER (Entrée) 10 RUN/HALT PROGRAM (Lancer/Interrompre le programme) 3 STOP (Arrêt) 7 BACK (Retour) 4 Diode électroluminescente 8 FLECHES 11 MANUAL OPERATION (Fonctionnement manuel) Fonctionnement Configuration des paramètres système 1. Appuyez sur MENU. 2. Accédez à SYSTEM SETUP (Configuration système). 3. Modifiez les options applicables. Option Description Réglage heure/date Permet de définir l'heure (format 24 h) et la date. Communication Permet de sélectionner la vitesse en baud (19 200, 38 400, 57 600 ou 115 200) et le protocole (Modbus RTU ou ASCII) pour le port série. Configuration de la base Permet de sélectionner la base d'échantillonneur (réfrigéré tout climat, réfrigéré, portable standard ou portable compact). Langue Permet de sélectionner la langue du transmetteur. Et capteur liq Permet d'étalonner le capteur liquide ou de sélectionner les valeurs par défaut. Régler le contraste Permet de régler le contraste de l'écran LCD. Cycle de fonctionnement de la pompe Permet de modifier le cycle de fonctionnement de la pompe. Plage : de 50 à 100 % (100 % par défaut). 42 Français Option Description Configuration du mot de passe Permet de configurer un mot de passe pour contrôler l'accès aux menus de configuration du programme et de configuration système. Sélectionnez ENABLE (Activer) > YES (Oui) pour définir un nouveau mot de passe ou ENABLE (Activer) > NO (Non) pour activer le mot de passe existant. Si le mot de passe est défini pour la première fois, utilisez 900900 en tant que mot de passe actuel. Durée de vie de la tuyauterie Permet de définir un rappel lorsque la tuyauterie de la pompe doit être remplacée. Accédez à ENABLE (Activer) > RESET CYCLES (Réinitialiser cycles) pour réinitialiser le nombre de cycles de la pompe sur 0. Accédez à ENABLE (Activer) > CYCLE LIMIT (Limite de cycle) pour modifier le nombre de cycles de la pompe utilisé pour le rappel. Accéder à ENABLE (Activer) > STATUS (Etat) pour afficher le nombre actuel de cycles de la pompe. SDI-12 (en option) Permet de configurer le fonctionnement d'un appareil SDI-12. Cette option s'affiche uniquement lorsqu'un appareil est détecté. Configuration d'un programme d'échantillonnage Un programme d'échantillonnage comprend tous les paramètres nécessaires pour prélever automatiquement des échantillons. Prélevez les échantillons à intervalles réguliers ou lorsque le programme d'échantillonnage est terminé. 1. Appuyez sur MENU. 2. Accédez à PROGRAM SETUP (Configuration du programme) > MODIFY ALL (Modifier tout). Le premier écran s'affiche. 3. Sélectionnez ou saisissez les valeurs pour chaque paramètre. Utilisez les touches fléchées gauche et droite pour déplacer le curseur à gauche ou à droite. Utilisez les touches fléchées haut et bas pour changer la valeur. Changez les unités au besoin (gal ou ml) avec les touches fléchées. Option Description Bouteilles Quantité de bouteilles : le nombre de bouteilles dans l'échantillonneur (1, 2, 4, 8 ou 24). Fermeture de bouteille pleine : activé/désactivé. Si activé, le transmetteur vérifie les conditions de débordement de la bouteille. Volume de la bouteille : la capacité de chaque bouteille. Plage : 50 à 65 000 ml (0,01 à 17,17 gal) Tuyauterie d'admission Longueur de la tuyauterie : la longueur du tube d'admission depuis la crépine jusqu'au capteur de liquide. Plage : 100 à 3 000 cm ou 3 à 99 pieds. Une longueur précise est nécessaire pour obtenir un volume d'échantillon précis. Type de tube d'admission : le diamètre et le matériau du tube d'admission (¼ po. vinyle, 3/8 po. vinyle ou 3/8 po. Teflon). Retard de programme Activer/désactiver : une fois activé, le programme d'échantillonnage démarre à une heure ou un nombre de comptages spécifié. Date et heure : la date et l'heure de démarrage du programme (format 24 h). Comptages : le nombre de comptages ou de pulsations d'un débitmètre lorsque le programme démarre (1 à 9 999 comptages). Si le rythme de l'échantillon est défini ultérieurement sur l'heure, le retard de programme sera désactivé. Français 43 Option Description Rythme/collecte d'échantillon Remarque : Lorsque CTVV est utilisé : Basés sur le temps—Reportez-vous à Collecte basée sur le temps à la page 44. • • • • • Le capteur de liquide est toujours activé. La distribution d'échantillon est composite. Le mode d'exécution est non continu. Le premier échantillon est pris après l'intervalle. La source du débit est toujours l'entrée 4–20 mA Basés sur le débit—Reportez-vous à Collecte basée sur le débit à la page 45. Distribution d'échantillon Fournir des échantillons à tous > YES (Oui) Chaque échantillon sera livré dans toutes les bouteilles. Terminer après le dernier : le programme s'interrompt après le dernier échantillon (1 à 999). Continu : le programme continue jusqu'à ce qu'il soit interrompu manuellement. Fournir des échantillons à tous > NO (Non) Les échantillons sont livrés à un sous-réseau de bouteilles . Echantillons/bouteille : nombre d'échantillons à prélever dans chaque bouteille (1 à 999). Bouteilles/échantillon : nombre de bouteilles qui contiendront le même échantillon. Capteur de liquide Activé ou désactivé. Si activé, le volume d'échantillon doit être calibré selon le temps. Sample volume (Volume d’échantillon) Le volume de chaque échantillon en ml (1 gal = 3 785,4 ml). Si le mode bouteilles/échantillon est sélectionné, chaque bouteille reçoit un volume entier d'échantillon (10 à 10 000 ml). Assurez-vous que le volume d'échantillon ne dépasse pas le volume de bouteille. Les volumes d'échantillon sont arrondis aux 10 ml les plus proches. Rinçages d'admission Le nombre de rinçages du tube d'admission avant prélèvement d'un échantillon (0 à 3). Tentatives d'échantillonnage Le nombre d'essais d'échantillonnage après échec d'un échantillonnage (0 à 3). ID site Le nom d'un emplacement d'échantillonnage (jusqu'à 12 caractères). L'ID du site est utilisé comme le nom prédéfini si le programme est enregistré. Échantillonnage avancé Permet d'accéder au menu d'échantillonnage avancé. Si nécessaire, configurez les options dans le menu d'échantillonnage avancé. Lancement du programme La création du programme d'échantillonnage basique se termine. L'échantillonneur invite l'utilisateur à démarrer ou annuler le programme. Collecte basée sur le temps 1. Sélectionnez ou saisissez les valeurs pour chaque option. Option Description Intervalle de rythme Prélève les échantillons à intervalles réguliers, en heures et minutes (0:01–999:00). 44 Français Option Description Prendre le premier échantillon Le programme démarre immédiatement ou après le premier intervalle. Volume variable Indique si les volumes d'échantillon varient (en fonction du débit). Si oui, reportez-vous à Utiliser le volume variable à la page 45. Si non, reportez-vous à Ne pas utiliser de volume variable à la page 45. Utiliser le volume variable Le rythme d'échantillonnage est volume variable à intervalle fixe (CTVV). 1. Sélectionnez ou saisissez les valeurs pour chaque option. Option Descriptions Sélectionner l'unité Sélectionnez les unités de débit utilisées. Définir l'entrée 4 mA Saisissez le débit qui correspond à l'entrée 4 mA Définir l'entrée 20 mA Saisissez le débit qui correspond à l'entrée 20 mA Débit moyen Indiquez le débit moyen pour le site (1–999 999). Les unités sont indiquées par les unités sélectionnées ci-dessus. Volume désiré pour la durée Indiquez le volume total devant être collecté lors du programme d'échantillonnage (10–10 000 ml). Période de collecte Indiquez la période de collecte totale en heures et minutes (0:01–999:00). Volume minimum d'échantillon Si le volume de débit n'est pas suffisant pour collecter ce volume minimum, l'échantillon est ignoré (10–10 000 ml). Ne pas utiliser de volume variable 1. Sélectionnez ou saisissez les valeurs pour l'option. Option Description Prendre le premier échantillon Vous permet de sélectionner si le programme commence immédiatement ou à la fin du premier intervalle. Collecte basée sur le débit Spécifiez la source du débit pour l'échantillonnage basé sur le débit : • Comptages—Reportez-vous à Utiliser les comptages pour la source de débit à la page 45. • Entrée 4–20 mA—Reportez-vous à Utilisez l'entrée de 4–20 mA pour la source de débit à la page 46. Utiliser les comptages pour la source de débit 1. Sélectionnez ou saisissez les valeurs pour chaque option. Option Description Prendre l'échantillon chaque Prélève des échantillons à intervalles de débit réguliers, en comptages (1 à 9 999 comptages). Durée de temporisation Prélève un échantillon si le volume de débit est très faible. Français 45 Option Description Durée Durée maximale entre les échantillons (0:01–999:00). Le retardateur démarre à nouveau après chaque prélèvement d'échantillon. Prendre le premier échantillon Le programme démarre immédiatement ou après le premier intervalle de temps. Utilisez l'entrée de 4–20 mA pour la source de débit 1. Sélectionnez ou saisissez les valeurs pour chaque option. Option Descriptions Sélectionner l'unité Permet de sélectionner les unités de débit utilisées. Définir l'entrée 4 mA Saisissez le débit qui correspondrait à l'entrée 4 mA. Définir l'entrée 20 mA Saisissez le débit qui correspondrait à l'entrée 20 mA. Prendre l'échantillon chaque Saisissez le débit total après lequel l'échantillon doit être tiré. Temps de dépassement Sélectionnez Activé pour forcer la collecte d'un échantillon si le volume de débit est exceptionnellement bas. Durée Durée maximale entre les échantillons (0:01–999:00). Le retardateur démarre à nouveau après chaque prélèvement d'échantillon. Volume d’échantillon Entrez le volume à collecter par échantillon. Si le mode bouteilles/échantillon est sélectionné, chaque bouteille reçoit un volume entier d'échantillon (100 à 10 000 ml). Cette option est ignorée si le volume variable est sélectionné. Restauration des paramètres par défaut L'option de restauration définit tous les paramètres du programme sur les paramètres par défaut. 1. Appuyez sur MENU. 2. Accédez à PROGRAM SETUP (Configuration du programme) > RESTORE (Restaurer). 3. Sélectionnez YES (Oui). Lancer ou arrêter un programme Utilisez la touche RUN/HALT PROGRAM (Lancer/Interrompre le programme) pour lancer ou interrompre un programme. Jusqu'à 12 heures/dates de démarrage/arrêt peuvent être définis par l'utilisateur. Un programme doit être arrêté avant que la configuration du programme ou du système puisse être modifiée. 1. Appuyez sur RUN/HALT PROGRAM (Lancer/Interrompre le programme). 2. Sélectionnez l’une des options. Option Description Début Démarrez le programme acutellement chargé. L'état change en RUNNING (En service). Halt (Interrompre) Interrompt le programme temporairement. L'état change en HALTED (Interrompu). Reprendre : Le programme reprend à partir de l'endroit auquel il a été interrompu. L'état change en RUNNING (En service). Start over (Reprendre) Le programme reprend depuis le début. L'état change en RUNNING (En service). End program (Terminer le programme) Termine le programme en cours. L'état change en COMPLETE (Terminé). 46 Français Fonctionnement manuel Utilisez le fonctionnement manuel pour collecter un échantillon instantané, déplacer le bras du distributeur ou actionner la pompe. Pour plus d'informations, reportez-vous à la version enrichie de ce manuel. Calibrage de volume Le volume d'échantillon peut être étalonné automatiquement avec le capteur de liquide ou manuellement. Si le capteur de liquide est utilisé pour le calibrage de volume, le capteur est ajusté pour mesurer avec précision tous les volumes programmés. Si le capteur de liquide est désactivé, le volume pour les échantillons, les rinçages et l'eau de précipitation doit être étalonné manuellement. Pour des applications typiques, l'eau du robinet peut être utilisée pour l'étalonnage de volume. Si la composition d'un échantillon est sensiblement différente des échantillons d'eau typiques, étalonnez l'échantillonneur en utilisant le liquide à prélever. Tous les programmes doivent être interrompus avant l'étalonnage. Vous permet de calibrer le capteur de liquide La précision du capteur de liquide peut changer en fonction du type de liquide en cours d'échantillonnage. Par exemple, le volume d'un échantillon fortement turbide peut ne pas être aussi précis que celui d'une eau claire. Le capteur de liquide peut être étalonné avec l'eau d'échantillon afin d'améliorer la précision du volume. 1. Placez le tube d'admission dans l'eau d'échantillon. 2. Appuyez sur MENU. 3. Accédez à PROGRAM SETUP (Configuration du programme) > MODIFY SELECTED (Modifier la sélection) > LIQUID SENSOR (Capteur de liquide) > ENABLE (Activer) pour vous assurer que le capteur de liquide est activé. 4. Accédez à SYSTEM SETUP (Configuration du système) > LIQ SENSOR CAL. (Etal. du capteur de liquide). 5. Sélectionnez PERFORM CAL. (Effectuer étal.). 6. Sélectionnez START (Démarrer). La pompe fonctionne en sens inverse pour purger le tube d'admission. Elle fonctionne alors en avant. 7. Vérifiez le flux d'échantillon dans le tube d'admission afin de vous assurer que l'échantillon traverse le capteur de liquide. 8. Lorsque l'échantillon traverse le capteur de liquide et qu'il n'y a aucune bulle, appuyez sur la touche STOP (Arrêter). Le tube d'admission est automatiquement purgé. L'étalonnage du capteur de liquide est terminé. 9. Sélectionnez DONE (Terminé) pour quitter ou REPEAT (Répéter) pour effectuer à nouveau l'étalonnage. Restauration de l'étalonnage d'usine L'étalonnage de volume pour le capteur de liquide peut être réinitialisé aux paramètres par défaut. 1. Sélectionnez PROGRAM SETUP (Configuration du programme) > MODIFY SELECTED (Modifier la sélection) > LIQUID SENSOR (Capteur de liquide) > ENABLE (Activer) pour vous assurer que le capteur de liquide est activé. 2. Appuyez sur VOLUME CALIBRATION (Etalonnage du volume) et sélectionnez RESET CAL. (Réinitialiser Etal.). L'étalonnage du volume est défini sur les paramètres par défaut. Le message « value has been reset to 0 » (valeur réinitialisée sur 0) s'affiche. Etalonnage du volume d'échantillon avec le capteur de liquide Etalonnez le volume d'échantillon avec le capteur de liquide activé pour ajuster les volumes d'échantillon avec précision. Français 47 1. Accédez à PROGRAM SETUP (Configuration du programme) > MODIFY SELECTED (Modifier la sélection) > LIQUID SENSOR (Capteur de liquide) > ENABLE (Activer) pour vous assurer que le capteur de liquide est activé. 2. Placez le tube d'admission dans l'eau d'échantillon ou de l'eau du robinet. Remarque : utilisez l'eau d'échantillon pour une plus grande précision. 3. Placez le tuyau d'évacuation de la pompe dans un cylindre gradué. 4. Appuyez sur VOLUME CALIBRATION (Etalonnage du volume) et sélectionnez CALIBRATION (Etalonnage). 5. Sélectionnez START (Démarrer). La pompe fonctionne en sens inverse pour purger le tube d'admission. Elle fonctionne alors en avant et collecte le volume d'échantillon spécifié dans le programme d'échantillonnage. La pompe fonctionne en sens inverse pour purger le tube d'admission. 6. Lorsque l'échantillon a été collecté, sélectionnez DONE (Terminé). Comparez le volume d'échantillon dans le cylindre gradué avec le volume d'échantillon saisi dans le menu de configuration du programme. 7. Si le volume qui a été collecté est différent du volume d'échantillon dans le programme en cours, entrez le volume qui a été réellement collecté. Appuyez sur ENTER. Le capteur est ajusté pour mesurer avec précision tous les volumes programmés. 8. Sélectionnez START (Démarrer) pour lancer le programme d'échantillonnage ou CANCEL (Annuler) pour quitter le menu d'étalonnage du volume. Etalonnage manuel du volume de l'échantillon Lorsque le capteur de liquide est désactivé, tous les volumes doivent être étalonnés manuellement en fonction de la durée. Le volume d'échantillon est calibré pour le volume indiqué dans le programme en cours. Si le volume d'échantillon est modifié dans le programme, le volume d'échantillon doit être à nouveau étalonné pour le nouveau volume. 1. Sélectionnez PROGRAM SETUP (Configuration du programme) > MODIFY SELECTED (Modifier la sélection) > LIQUID SENSOR (Capteur de liquide) > DISABLE (Désactiver) pour vous assurer que le capteur de liquide est désactivé. 2. Placez le tube d'admission dans l'eau d'échantillon ou de l'eau du robinet. 3. Placez le tuyau d'évacuation de la pompe dans un cylindre gradué. 4. Appuyez sur VOLUME CALIBRATION (Etalonnage du volume). Si plusieurs volumes sont affichés, sélectionnez l'un d'eux pour l'étalonnage. 5. Sélectionnez START (Démarrer). La pompe fonctionne en sens inverse pour purger le tube d'admission. Elle fonctionne alors en avant et lance le prélèvement de l'échantillon. 6. Arrêtez la pompe au volume choisi ou rincez l'endroit : • Volume d'échantillon : appuyez sur STOP (Arrêter) lorsque le volume spécifié dans le programme principal est prélevé. • Rincer : appuyez sur STOP (Arrêter) lorsque le liquide dans le tube d'admission atteint le détecteur de liquide. • Volume d'eaux de précipitation : appuyez sur STOP (Arrêter) lorsque le volume spécifié dans le programme d'eaux de précipitation est prélevé. 7. Sélectionnez DONE (Terminé) pour accéder au prochain étalonnage de volume ou REPEAT (Répéter) pour effectuer à nouveau l'étalonnage. 8. Lorsque tous les volumes ont été étalonnés, sélectionnez DONE (Terminé). 9. Reliez le tube d'évacuation au raccord de tube sur l'échantillonneur. 10. Sélectionnez START (Démarrer) pour lancer le programme d'échantillonnage ou CANCEL (Annuler) pour quitter. 48 Français Vérification du volume d'échantillon Pour vous assurer que le volume d'échantillon est précis, prenez un échantillon instantané. Ne revenez pas dans le calibrage pour vérifier le volume, puisque la compensation de volume est remise à zéro au début d'un calibrage. 1. 2. 3. 4. 5. 6. 7. 8. Appuyez sur MANUAL OPERATION (Fonctionnement manuel). Sélectionnez GRAB SAMPLE (Echantillon instantané. Placez le tube d'admission dans l'eau d'échantillon. Placez le tuyau d'évacuation de la pompe dans un cylindre gradué. Si le capteur de liquide est activé, saisissez le volume à vérifier. Si le capteur de liquide est désactivé, entrez le volume indiqué dans le programme. Appuyez sur ENTER. Le cycle de pompe démarre (purge-échantillon-purge). Comparez le volume qui a été prélevé dans le cylindre gradué au volume attendu. Si le volume prélevé est incorrect, effectuez à nouveau l'étalonnage. Maintenance DANGER Dangers multiples. Seul le personnel qualifié doit effectuer les tâches détaillées dans cette section du document. DANGER Risque d'électrocution Coupez l'alimentation de l'instrument avant d'effectuer des activités de maintenance ou d'entretien. AVERTISSEMENT Exposition aux risques biologiques. Respectez des protocoles de manipulation sécurisés lorsque vous êtes en contact avec les flacons d'échantillon et les composants de l'échantillonneur. AVERTISSEMENT Dangers multiples. Le technicien doit s'assurer que l'équipement fonctionne correctement et de manière sécurisée après avoir exécuté une procédure de maintenance. Nettoyage de l'instrument ATTENTION Risque d’incendie. Ne pas utiliser d'agents inflammables pour nettoyer l'appareil. AVIS Ne nettoyez pas le radiateur du compartiment du transmetteur avec des liquides, toutes sortes confondues. Si l'eau n'est pas suffisante pour nettoyer le transmetteur et la pompe, déconnectez le transmetteur et éloignez-le de l'échantillonneur. Laissez sécher le transmetteur et la pompe avant de réinstaller les pièces et de les remettre en service. Nettoyez l'échantillonneur comme suit : • Boîtier de l'échantillonneur : nettoyez les surfaces internes et externes avec un tissu humide et un détergent doux. N'utilisez pas de nettoyeurs abrasifs ou de dissolvants. Français 49 • Bouteilles d'échantillon : nettoyez la bouteille avec une brosse, de l'eau et un détergent doux. Rincez avec de l'eau douce, puis avec de l'eau distillée. Utilisez un autoclave pour nettoyer les bouteilles en verre si nécessaire. Remplacement du dessiccateur Une cartouche déshydratante dans le transmetteur absorbe l'humidité et empêche la corrosion. Surveillez la couleur du déshydratant à travers la fenêtre (Figure 12). Un déshydratant non utilisé est jaune. Lorsque la couleur devient verte, remplacez le déshydradant. 1. Desserrez et retirez la cartouche déshydratante (Figure 12). Retirez le bouchon et le déshydratant utilisé. Remplissez le tube du dessiccant avec un dessiccant frais. Placez le bouchon. Mettez de la graisse sur le joint torique. Installez le tube du déshydratant dans le transmetteur. 2. 3. 4. 5. 6. Figure 12 Cartouche desséchante 1 Prise 3 Joint torique 2 Ensemble du 4 Fenêtre du dessiccant Maintenance de la pompe ATTENTION Risque de pincement. Coupez l'alimentation de l'instrument avant d'effectuer des activités de maintenance ou d'entretien. Remplacement du tuyau de la pompe AVIS L'utilisation d'une tuyauterie autre que celle fournie par le fabricant peut causer une usure excessive des pièces mécaniques et/ou des performance moindres de la pompe. Recherchez sur la tuyauterie de la pompe des traces d'usure au niveau de la zone de friction des rouleaux avec la tuyauterie. Remplacez la tuyauterie lorsque le message de rappel s'affiche ou lorsque la tuyauterie montre des signes d'usure. Prérequis : • Tuyauterie de la pompe – prédécoupée ou complète, 4,6 m ou 15,2 m (15 ou 50 pieds) 1. Déconnectez l'alimentation du transmetteur. 2. Si la tuyauterie complète est utilisée, coupez-la et ajoutez des points d'alignement. Consultez le manuel complet sur le site Web du fabricant. 3. Retirez la tuyauterie de la pompe tel qu'illustré dans les étapes suivantes. 50 Français 4. Nettoyez les résidus de silicone à l'intérieur du corps de la pompe et sur les rouleaux. 5. Installez la nouvelle tuyauterie de la pompe tel qu'illustré dans les étapes suivantes. Français 51 Nettoyage du rotor Nettoyez le rotor, les voies de tube de pompe et le boîtier de pompe avec un détergent doux. Reportez-vous au chapitre Remplacement du tuyau de la pompe à la page 50 et aux étapes illustrées suivantes. Remplacement du tube du bras de distribution Le bras du distributeur se déplace au-dessus de chaque bouteille pendant l'échantillonnage des bouteilles multiples. Remplacez le tube dans le bras de distribution lorsque le tube est usé. Assurezvous que le tube correct est utilisé pour le distributeur ainsi que le bras de distributeur adéquats. Remarque : Le tube du distributeur n'est pas le même que celui de la pompe. Si vous installez le tube de la pompe dans le distributeur, ce dernier peut être endommagé. Les échantillons ne pourront pas être correctement prélevés car le bras du distributeur ne se déplacera pas aisément. 1. Retirez le tube du bras de distribution et du haut de la pièce centrale. 2. Insérez le nouveau tube dans le bras de distribution. Assurez-vous que le tube est aligné selon les dimensions spécifiées dansFigure 13. 3. Insérez l'autre extrémité du tube dans le raccord sur le haut de la pièce centrale. 4. Effectuez le test de diagnostic pour le distributeur afin de vous assurer que le système fonctionne correctement. Figure 13 Ensemble distributeur 1 Buse 3 Moteur de distributeur 2 Bras de distributeur 4 Axe 52 Français 5 Tube distributeur Remplacement des fusibles L'utilisateur ne peut pas remplacer les fusibles. Contactez l’assistance technique. Un fusible grillé peut indiquer que l'échantillonneurest défaillant et qu'une réparation est nécessaire. Français 53 Tabla de contenidos Especificaciones en la página 54 Funcionamiento en la página 68 Información general en la página 56 Mantenimiento en la página 75 Instalación en la página 59 Información adicional En el sitio web del fabricante encontrará información adicional. Especificaciones Las especificaciones están sujetas a cambios sin previo aviso. Muestreador portátil SD900 Especificación Base estándar Base compacta Base compuesta Dimensiones Diámetro: 50,5 cm (19,9 pulg.) Diámetro: 44,1 cm (17,4 pulg.) Diámetro: 50,28 cm (19,8 pulg.) Altura: 61 cm (24 pulg.) Altura: 79,75 cm (31,4 pulg.) 12,2 kg (27 lb) con botellas de polietileno de 575 ml (19,44 oz) (24x) 15 kg (36 lb) con botellas de vidrio de 950 ml (32,12 oz) (12x) Altura: 69,4 cm (27,3 pulg.) Peso 15 kg (35,6 lb) con botellas de polietileno de 1 l (24x) 14,8 kg (32,6 lb) con recipiente de polietileno de 10 l (2,5 gal) (1x) 12,9 kg (28,3 lb) con recipiente de polietileno de 10 l (2,5 gal) (1x) Carcasa ABS resistente a impactos, estructura de 3 piezas; base con pared doble con aislamiento de 2,54 cm (1 pulg.) Aislamiento: contacto directo de la botella con hielo. Temperatura de la muestra 0-60 °C (32-140 °F) Filtros De acero inoxidable 316 en tamaño estándar, de alta velocidad o bajo perfil para aplicaciones a poca profundidad y de Teflon® o acero inoxidable 316 en tamaño estándar. Tubos de entrada de muestras D.I. de 9,5 mm (3/8 pulg.), polietileno con revestimiento de vinilo o teflón Capacidad de la botella para muestras Botellas de polietileno de 1 l (0,26 gal) y/o botellas de vidrio de 350 ml (11,83 oz) (24x) Botellas de polietileno de 2,3 l (0,6 gal) y/o botellas de vidrio de 1,9 l (0,5 gal) (8x) Botellas de polietileno de 3,8 l (1 gal) y/o botellas de vidrio de 3,8 l (1 gal) (4x) Botellas de polietileno de 3,8 l (1 gal) y/o botellas de vidrio de 3,8 l (1 gal) (2x) Recipiente de polietileno compuesto de 21 l (5,5 gal) o 15-L (4 gal), de polietileno de 20 l (5,25 gal), de polietileno de 10 l (2,5 gal) o de vidrio de 10 l (2,5 gal) (1x) 54 Español Botellas de polietileno de 575 ml (19,44 oz) (24x) Botellas de vidrio de 950 ml (32,12 oz) (8x) Botella de polietileno de 10 l (2,5 gal) (1x) Botella de vidrio de 10 l (2,5 gal) (1x) Botella de polietileno de 21 l (5,5 gal) (1x) Controlador SD Especificación Detalles Dimensiones (an. x alt. x prof.) 29,2 x 17,1 x 26,4 cm (11½ x 6¾ x 103/8 pulg.) Peso 4.2 kg (9.26 libras) Carcasa Mezcla de PC/ABS, NEMA 4X, 6, IP 67 Requisitos de alimentación 15 V CC suministrados por una fuente de alimentación 8754500; 15 V CC suministrados por una fuente de alimentación integrada Protección contra sobrecarga 7 A, fusible de la línea CC para la bomba Temperatura de funcionamiento 0 a 50 °C (32 a 122 °F) Temperatura de almacenamiento –30 a 60 °C (–22 a 140 °F) Humedad en almacenamiento/funcionamiento Condensación 100% Bomba Peristáltica de alta velocidad, con rodillos de nilatron montados sobre resortes Recinto de la bomba IP37 Tubería de la bomba 9,5 mm de diámetro interior x 15,9 mm de diámetro exterior (3/8 pulg. de diámetro interior x 5/8 pulg. de diámetro exterior) de silicona Vida de la tubería de la bomba Ciclos de 20.000 muestras con: volumen de muestras de 1 l (0,3 galones), 1 enjuague, intervalo de cadenciado de 6 minutos, 4,9 m (16 pies) del tubo de entrada de 3/8 pulg. , 4,6 m (15 pies) de elevación vertical, temperatura de la muestra de 21 °C (70 °F) Elevación de la muestra vertical 8,5 m (28 pies) máximo para: 8,8 m (29 pies) del tubo de entrada de vinilo de 3/8 pulg. a nivel del mar a 20–25 °C (68–77 °F) Caudal de la bomba 4,8 l/min (1,25 gpm) a 1 m (3 pies) de elevación vertical con un tubo de entrada de 3/8 pulg. típico Volumen de la muestra Se puede programa en incrementos de 10 ml (0,34 onzas) desde 10 hasta 10.000 ml (3,38 onzas a 2,6 galones) Repetibilidad del volumen de la muestra (típica) ±5% del volumen de la muestra de 200 ml usando detección de líquido sin calibrar con: elevación vertical de 4,6 m (15 pies), 4,9 m (16 pies) de un tubo de entrada de vinilo de 3/8 pulg. , una sola botella, dispositivo de corte para botella llena a temperatura ambiente y elevación de 1524 m (5000 pies) Precisión del volumen de la muestra (típico) ±10% del volumen de la muestra de 200 ml usando detección de líquido sin calibrar con: elevación vertical de 4,6 m (15 pies), 4,9 m (16 pies) de un tubo de entrada de vinilo de 3/8 pulg. , una sola botella, dispositivo de corte para botella llena a temperatura ambiente y elevación de 1524 m (5000 pies) Modos de muestreo Intervalos: de tiempo fijo, de caudal fijo, de tiempo variable, de caudal variable y de caudal variable en tiempo constante (CTVV, por sus siglas en inglés). Consulte la información sobre muestras de caudal variable en tiempo constante (CTVV) en la versión ampliada de este manual. Distribución: compuesta en una sola botella, compuesta en varias botellas, aislada en varias botellas, botellas por muestra, muestras por botella, combinación de botellas por muestra, muestras por botella Modos de ejecución Continuos o no continuos, con cantidad de muestras ingresada por el usuario Español 55 Especificación Detalles Múltiples programas Almacena hasta tres programas de muestreo. Velocidad de transferencia (típica) 0,9 m/s (2,9 pies/s) con: elevación vertical de 4,6 m (15 pies), 4,9 m (16 pies) de tubos de entrada de vinilo de 3/8 pulg. , 21 °C (70 °F) y elevación de 1524 m (5000 pies) Sensor de líquido Ultrasónico. Cuerpo: Ultem aprobado según norma NSF ANSI 51, cumple con USP Clase VI Histórico de muestreo Hasta 510 registros Purga de aire Aire purgado automáticamente antes y después de cada muestra. La duración compensa automáticamente las variaciones en las longitudes del tubo de entrada. Registro de eventos 510 registros Conexiones Alimentación, auxiliar, comunicaciones serie, distribuidor, SDI-12 Materiales impregnados Acero inoxidable, polietileno, teflón, Ultem, silicona Comunicaciones RS232, Modbus, SDI-12 Garantía 1 año ® Información general En ningún caso el fabricante será responsable de ningún daño directo, indirecto, especial, accidental o resultante de un defecto u omisión en este manual. El fabricante se reserva el derecho a modificar este manual y los productos que describen en cualquier momento, sin aviso ni obligación. Las ediciones revisadas se encuentran en la página web del fabricante. Información de seguridad AVISO El fabricante no es responsable de ningún daño debo a un mal uso de este producto incluyendo, sin limitación, daños directos, fortuitos o circunstanciales y reclamaciones sobre los daños que no estén recogidos en la legislación vigente. El usuario es el responsable de la identificación de los riesgos críticos y de tener los mecanismos adecuados de protección de los procesos en caso de un posible mal funcionamiento del equipo. Lea todo el manual antes de desembalar, instalar o trabajar con este equipo. Ponga atención a todas las advertencias y avisos de peligro. El no hacerlo puede provocar heridas graves al usuario o daños al equipo. Asegúrese de que la protección proporcionada por el equipo no está dañada. No utilice ni instale este equipo de manera distinta a lo especificado en este manual. Uso de la información sobre riesgos PELIGRO Indica una situación potencial o de riesgo inminente que, de no evitarse, provocará la muerte o lesiones graves. ADVERTENCIA Indica una situación potencial o inminentemente peligrosa que, de no evitarse, podría provocar la muerte o lesiones graves. PRECAUCIÓN Indica una situación potencialmente peligrosa que podría provocar una lesión menor o moderada. AVISO Indica una situación que, si no se evita, puede provocar daños en el instrumento. Información que requiere especial énfasis. 56 Español Etiquetas de precaución Lea todas las etiquetas y rótulos adheridos al instrumento. En caso contrario, podrían producirse heridas personales o daños en el instrumento. El símbolo que aparezca en el instrumento se comentará en el manual con una declaración de precaución. Este es un símbolo de alerta de seguridad. Obedezca todos los mensajes de seguridad que se muestran junto con este símbolo para evitar posibles lesiones. Si se encuentran sobre el instrumento, consulte el manual de instrucciones para obtener información de funcionamiento o seguridad. Este símbolo indica un peligro de pellizco potencial. En Europa, el equipo eléctrico marcado con este símbolo no se debe desechar mediante el servicio de recogida de basura doméstica o pública. Devuelva los equipos viejos o que hayan alcanzado el término de su vida útil al fabricante para su eliminación sin cargo para el usuario. Certificación Reglamentación canadiense sobre equipos que provocan interferencia, IECS-003, Clase A Registros de pruebas de control del fabricante. Este aparato digital de clase A cumple con todos los requerimientos de las reglamentaciones canadienses para equipos que producen interferencias. Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences. FCC Parte 15, Límites Clase "A" Registros de pruebas de control del fabricante. Este dispositivo cumple con la Parte 15 de las normas de la FCC estadounidense. Su operación está sujeta a las siguientes dos condiciones: 1. El equipo no puede causar interferencias perjudiciales. 2. Este equipo debe aceptar cualquier interferencia recibida, incluyendo las interferencias que pueden causar un funcionamiento no deseado. Los cambios o modificaciones a este equipo que no hayan sido aprobados por la parte responsable podrían anular el permiso del usuario para operar el equipo. Este equipo ha sido probado y encontrado que cumple con los límites para un dispositivo digital Clase A, de acuerdo con la Parte 15 de las Reglas FCC. Estos límites están diseñados para proporcionar una protección razonable contra las interferencias perjudiciales cuando el equipo está operando en un entorno comercial. Este equipo genera, utiliza y puede irradiar energía de radio frecuencia, y si no es instalado y utilizado de acuerdo con el manual de instrucciones, puede causar una interferencia dañina a las radio comunicaciones. La operación de este equipo en un área residencial es probable que produzca interferencia dañina, en cuyo caso el usuario será requerido para corregir la interferencia bajo su propio cargo. Pueden utilizarse las siguientes técnicas para reducir los problemas de interferencia: 1. Desconecte el equipo de su fuente de alimentación para verificar si éste es o no la fuente de la interferencia. 2. Si el equipo está conectado a la misma toma eléctrica que el dispositivo que experimenta la interferencia, conecte el equipo a otra toma eléctrica. 3. Aleje el equipo del dispositivo que está recibiendo la interferencia. 4. Cambie la posición de la antena del dispositivo que recibe la interferencia. 5. Trate combinaciones de las opciones descritas. Español 57 Descripción general del producto PELIGRO Peligro químico o biológico. Si este instrumento se usa para controlar un proceso de tratamiento y/o un sistema de suministro químico para el que existan límites normativos y requisitos de control relacionados con la salud pública, la seguridad pública, la fabricación o procesamiento de alimentos o bebidas, es responsabilidad del usuario de este instrumento conocer y cumplir toda normativa aplicable y disponer de mecanismos adecuados y suficientes que satisfagan las normativas vigentes en caso de mal funcionamiento del equipo. PRECAUCIÓN Peligro de incendio. Este producto no ha so diseñado para utilizarse con líquidos inflamables. El muestreador portátil SD900 recoge muestras líquidas a intervalos específicos y conserva las muestras en botellas o recipientes. Utilice el muestreador para una amplia variedad de aplicaciones acuosas, así como para recoger contaminantes tóxicos y sólidos en suspensión. Configure el muestreador con diferentes dispositivos de retención, botellas o recipientes. Consulte Figura 1. Figura 1 Descripción general del producto 1 Base compacta 6 Conexiones del controlador 2 Base aislada estándar 7 Bomba 3 Sección central 8 Controlador 4 Fuente de alimentación 9 Sensor de líquido 5 Tapa superior 10 Base aislada estándar para recipiente de 21 l (5,5 gal) Componentes del producto Asegúrese de haber recibido todos los componentes. Consulte la Figura 2. Si faltan artículos o están dañados, póngase en contacto con el fabricante o el representante de ventas inmediatamente. 58 Español Figura 2 Componentes del producto 1 Base (estándar, compacta o compuesta) 6 Cargador de batería (opcional) 2 Componentes para la opción de botella única (la botella y el soporte se pueden cambiar) 7 Fuente de alimentación de CA (opcional) 3 Sección central con controlador 8 Filtro 4 Batería (opcional) 9 Tubo de entrada, revestimiento de vinilo o teflón 5 Tapa superior 10 Componentes para la opción de varias botellas (las botellas y los dispositivos de retención se pueden cambiar) Instalación PELIGRO Peligros diversos. Sólo el personal cualificado debe realizar las tareas descritas en esta sección del documento. Precauciones para espacios confinados PELIGRO Peligro de explosión. XXX La información que se incluye a continuación se ofrece para ayudar a los usuarios a comprender los peligros y riesgos asociados a los espacios confinados. El 15 de abril de 1993, el dictamen definitivo de la OSHA (Administración de Seguridad y Salud Ocupacional) sobre los Espacios Confinados que Requieren Permiso para Ingresar (CFR 1910.146), se hizo ley. Esta nueva norma afecta directamente a más de 250.000 sitios industriales de los Estados Unidos, y fue creada con el fin de proteger la salud y la seguridad de los trabajadores en espacios confinados. Definición de espacio confinado: Un espacio confinado es cualquier lugar o recinto que presente (o tenga la posibilidad inmediata de presentar) una o más de las siguientes condiciones: • Una atmósfera con una concentración de oxígeno que sea inferior al 19,5% o superior al 23,5% y/o una concentración de sulfuro de hidrógeno (H2S) superior a 10 ppm. Español 59 • Una atmósfera que pueda ser inflamable o explosiva debido a gases, vapores, nieblas, polvos o fibras. • Materiales tóxicos que, ante el contacto o la inhalación, puedan provocar lesiones, el deterioro de la salud o la muerte. Los espacios confinados no están destinados a ser ocupados por seres humanos. Los espacios confinados tienen entrada restringida y contienen riesgos conocidos o potenciales. Como ejemplos de espacios confinados encontramos las bocas de inspección, las chimeneas, los caños, las tinas, los armarios de distribución y demás lugares similares. Antes de entrar en espacios confinados y/o lugares con presencia de gases, vapores, nieblas, polvos o fibras peligrosos, se deben seguir siempre procedimientos de seguridad estándares. Antes de entrar en un espacio confinado, lea todos los procedimientos relacionados con la entrada a espacios confinados. Instalación mecánica Pautas para la instalación PELIGRO Peligro de explosión. El instrumento no está aprobado para su instalación en lugares peligrosos. Consulte en las instrucciones siguientes el modo de evaluar la ubicación. • Siga todas las precauciones de seguridad si el muestreador está instalado en un espacio reducido. Consulte Precauciones para espacios confinados en la página 59. • Asegúrese de que la temperatura del emplazamiento está dentro del rango de especificaciones. Consulte Especificaciones en la página 54. • Instale el muestreador en una superficie nivelada o cuelgue el muestreador con el arnés de suspensión, el soporte de arqueta o la barra de extensión. Consulte Instalación el muestreador en una boca de inspección en la página 60 y la documentación de instalación pertinente. • Lo más cerca posible de la fuente de muestras para reducir el retraso en el análisis. Consulte Conexión del muestreador en la página 63. • Para obtener información sobre la velocidad del transporte y la inclinación vertical máxima, consulte Especificaciones en la página 54. Instalación el muestreador en una boca de inspección Instale el muestreador por encima de la muestra de agua en una boca de inspección. Instale el muestreador con una barra de extensión o un soporte de arqueta. Instale la barra de extensión en la boca de inspección. La barra de extensión se sujeta por la presión que se ejerce contra las paredes. El soporte de arqueta tiene la misma anchura que la cubierta de la boca de inspección. Instale el soporte de arqueta directamente por debajo de la cubierta a modo de apoyo.Consulte el manual completo en la página web del fabricante.Consulte la documentación suministrada con los accesorios para instalar el muestreador. Instalación del distribuidor o dispositivo de corte por botella llena (opcional) El conjunto del distribuidor o dispositivo de corte por botella llena generalmente viene instalado de fábrica. Consulte la documentación del distribuidor o del dispositivo de corte por botella llena para obtener información sobre la instalación. Preparación del tomamuestras Limpieza de las botellas para muestras Limpie las botellas para muestras y los tapones con un cepillo, agua y detergente suave. Enjuague los recipientes con agua corriente y luego con agua destilada. 60 Español Instalación de una única botella Utilice una única botella para recoger la muestra compuesta. Cuando la botella está llena, el dispositivo de corte por botella llena detiene el programa de muestreo. 1. Limpie las botellas para muestras. Consulte la Limpieza de las botellas para muestras en la página 60. 2. Instale la botella para muestras como se indica en la Figura 3. Figura 3 Instalación de una única botella 1 Botella de polietileno, 10 l (2,5 gal) (1918) 6 Botella de polietileno, 21 l (5,5 gal) (6494) 2 Botella de vidrio, 10 l (2,5 gal) (6559) 7 Base compacta (8975) 3 Asistencia técnica (1502) 8 Base aislada estándar (8976) 4 Botella de polietileno, 15 l (4 gal) (1367) 9 Base aislada estándar para botella de 21 l (5,5 gal) (8561) 5 Botella de polietileno, 19 l (5 gal) (6498) Instalación de la primera botella para recogidas de muestras múltiples Utilice múltiples botellas para recoger muestras en botellas distintas o en más de una botella. El distribuidor mueve el tubo para muestras por encima de cada botella. Instale las botellas en la base del muestreador, tal como se muestra en la Figura 4. Instale la primera botella para muestras (número 1) por debajo de la etiqueta que está en la base del muestreador. Instale el resto de las botellas siguiendo una numeración creciente, en la dirección indicada en la etiqueta. Consulte la Figura 5 en la página 63 para obtener un diagrama de los componentes requeridos. Español 61 Figura 4 Instalación de la botella número 1 1 Ubicación de la botella número 1 (base compacta) 6 Bandas elásticas 2 Ubicación de la botella número 1 para 24 botellas 7 Base aislada estándar para botella de 21 l (5,5 gal) 3 Ubicación de la botella número 1 para 8 botellas 8 Base aislada estándar 4 Ubicación de la botella número 1 para 2 o 4 botellas 9 Base compacta 5 Dispositivo de retención Instalación de varias botellas Cuando se instalan varias botellas, un brazo del distribuidor mueve el tubo para muestras por encima de cada botella. La recogida de muestras se detiene automáticamente cuando ya se ha recogido el número especificado de muestras. 1. Limpie las botellas para muestras. Consulte la Limpieza de las botellas para muestras en la página 60. 2. Instale la primera botella para muestras (número 1) por debajo de la etiqueta que está en la base del muestreador. Consulte Instalación de la primera botella para recogidas de muestras múltiples en la página 61. 3. Coloque las botellas para muestras como se indica en la Figura 5. Si hay ocho botellas o más, compruebe que la primera botella está junto al indicador de primera botella y que se colocan en el sentido de las agujas del reloj. 4. Coloque el conjunto de botellas en el tomamuestras. Si hay ocho botellas o más, alinee los cables en las ranuras de la bandeja inferior. 62 Español Figura 5 Instalación para múltiples botellas 1 Regulador de posición/dispositivo de retención (2347) 8 Botellas de vidrio, 3,8 l (1 gal) (4x) (2216) o botellas de polietileno, 3,8 l (1 gal) (4x) (2217) 2 Dispositivo de retención (1422) 9 Botellas de vidrio, 1,9 l (0,5 gal) (8x) (1118) o botellas de polietileno, 2,3 l (0,61 gal) (8x) (657) 3 Dispositivo de retención (2190) 10 Botellas de polietileno, 1 l (0,26 gal) (24x) (737) 4 Dispositivo de retención (2189) 11 Botellas de vidrio, 350 ml (11,83 oz) (24x) (732) 5 Botellas de vidrio, 950 ml (32,12 oz) (8x) (2348) 12 Base compacta (8975) 6 Botellas de polietileno, 575 ml (19,44 oz) (24x) (1369) 13 Base estándar (8976) 7 Botellas de vidrio, 3,8 l (1 gal) (2x) (2214) o botellas de polietileno, 3,8 l (1 gal) (2x) (2215) Conexión del muestreador Instale el filtro en el medio de la corriente de la muestra (no cerca de la superficie o del fondo) para asegurarse de que recoge una muestra representativa. Consulte Figura 6. Consulte la Figura 7 para instalar el tubo de entrada. 1. Conecte los tubos al tomamuestras como se indica en la Figura 7. Nota: Si usa tubos con revestimiento de teflón, emplee el kit de conexión 2186. 2. Coloque el tubo de entrada y el filtro en la corriente principal de la fuente de origen de la muestra, donde el agua es turbulenta y está bien mezclada. • El tubo de entrada debe ser lo más corto posible. • Mantenga el tubo de entrada en pendiente vertical, para que se vacíe completamente entre muestra y muestra. Nota: Si no se puede colocar en pendiente vertical o si el tubo se encuentra bajo presión, desactive el sensor de líquido. Calibre el volumen de la muestra de forma manual. Consulte Calibración del volumen de la muestra de forma manual en la página 74. • Compruebe que el tubo de entrada no esté aplastado. Español 63 Figura 6 Instalación del instrumento 1 Filtro 3 Tubo de entrada 2 Elevación vertical Figura 7 Instalación del tubo de entrada Instalación eléctrica Conexiones del controlador ADVERTENCIA Riesgo de descargas eléctricas. El equipo conectado de forma externa debe someterse a una evaluación estándar de seguridad aplicable. En la Figura 8 se muestran las conexiones que se pueden realizar en el controlador. 64 Español Figura 8 Conectores del controlador 1 Dispositivo auxiliar 4 Opción de dispositivo SDI-12 2 Fuente de alimentación 5 Dispositivo de corte para botella llena/distribuidor 3 Comunicaciones en serie Conexión de un caudalímetro Conecte un caudalímetro al controlador para iniciar o detener el tomamuestras cuando el flujo de la muestra descienda por debajo de un valor especificado o lo supere. Recopilación de elementos: • Cable completo multiuso para los caudalímetros Sigma (o cable semicompleto 980 para el caudalímetro modelo 980). • Repartidor opcional para realizar conexiones adicionales. Se pueden conectar en serie dos o más repartidores. 1. Conecte un extremo del cable al caudalímetro. En el caso del caudalímetro del modelo 980, consulte el manual del usuario del caudalímetro del modelo 980. 2. Conecte el otro extremo del cable al conector del dispositivo auxiliar en el controlador. Nota: Si el caudalímetro tiene un cable de 6 patillas, utilice el cable adaptador de 6 patillas a 7 patillas. Conexión de un PC o una red de comunicaciones Conecte un PC o una red de comunicaciones al controlador para transferir datos o configurar el tomamuestras. Recopilación de elementos: • Cable serie, RS232 de 7 patillas a DB-9 (consulte la Figura 9). Conexiones: B a 5 (puesta a tierra de señal); D a 3 (RCD); F a 2 (TXD); G (puesta a tierra). 1. Conecte un extremo del cable serie al dispositivo o red de comunicaciones. 2. Conecte el otro extremo del cable al conector de comunicaciones serie del controlador. Figura 9 Cable de comunicaciones serie Español 65 Conexión el muestreador a la alimentación eléctrica PELIGRO Peligro de electrocución. Si este equipo se usa en exteriores o en lugares potencialmente húmedos, debe utilizarse un disyuntor de interrupción de circuito por falla a tierra (GFCI/GFI) para conectar el equipo a la alimentación eléctrica. PELIGRO Peligro de incendio. Instale un disyuntor de 15 A en la línea de alimentación. Un disyuntor puede ser la desconexión de alimentación local, si se encuentra en las proximidades del equipo. PELIGRO Peligro de electrocución. Se requiere una conexión de toma a tierra (PE). ADVERTENCIA Peligro de electrocución. Asegúrese de que sea fácil acceder a la desconexión de alimentación local. ADVERTENCIA Riesgo de electrocución. El suministro de alimentación se puede recalentar cuando el tiempo entre ciclos de bombeo es muy corto. AVISO Asegúrese de que la duración del ciclo de bombeo no provoca la sobrecarga de la fuente de alimentación. Consulte Cálculo del tiempo de ciclo de la bomba en la página 67. Conecte el muestreador a una batería o a una toma de alimentación de CA. Consulte los pasos ilustrados que se muestran a continuación. 66 Español Cálculo del tiempo de ciclo de la bomba La bomba debe permanecer apagada durante un período mínimo de tiempo entre los ciclos de muestreo, para que la fuente de alimentación no se sobrecaliente. 1. Calcule la cantidad total de tiempo que la bomba funcionará sin interrupción durante un ciclo de muestreo. Incluya todas las etapas: purga previa, lavado de la línea de admisión, muestreo, reintentos de muestreo y purga posterior. 2. Determine la cantidad mínima de tiempo que la bomba debe estar apagada entre los ciclos de muestreo con ayuda de la Figura 10. 3. El cadenciado o el intervalo de tiempo del programa del tomamuestras debe hacer que la bomba se apague entre los ciclos de muestreo durante el período de tiempo mínimo correspondiente. Figura 10 Tiempos de encendido y apagado de la bomba a 50 °C (122 °F) 1 Tiempo (en minutos) durante los cuales la bomba debe permanecer apagada 2 Tiempo (en minutos) durante los cuales la bomba funciona sin interrupción 3 Ejemplo: si la bomba funciona sin interrupción durante 5 minutos, debe permanecer apagada 15 minutos. Conexión de la alimentación Pulse el botón de encendido para encender o apagar el dispositivo. Compruebe que el suministro de alimentación (alimentación de CA o pilas) se ha instalado correctamente. Interfaz del usuario y navegación El teclado del controlador se muestra en la Figura 11. Utilice la tecla MENU (Menú) para configurar los programas de muestreo, definir los ajustes del controlador o llevar a cabo las pruebas de diagnóstico. Utilice las teclas de flechas, ENTER (Intro) y BACK (Retroceso) para desplazarse por los menús, realizar selecciones e introducir valores. Mire si aparecen flechas en la parte inferior o en el lateral de la pantalla para saber si hay más opciones disponibles. Español 67 Figura 11 Teclado del controlador 1 POWER (Alimentación) 5 MENU (Menú) 9 STATUS (Estado) 2 VOLUME CALIBRATION (Calibración del volumen) 6 ENTER (Intro) 10 RUN/HALT PROGRAM (Ejecutar/Detener programa) 3 STOP (Parada) 7 BACK (Atrás) 4 LED 8 TECLAS DE FLECHA 11 MANUAL OPERATION (Funcionamiento manual) Funcionamiento Configuración de los ajustes del sistema 1. Pulse MENU (Menú). 2. Vaya a SYSTEM SETUP (Configuración del sistema). 3. Cambie las opciones correspondientes. Opción Descripción Time/date setup (Configurar hora/fecha) Ajuste la hora (formato de 24 horas) y la fecha. Communication (Comunicación) Seleccione la velocidad en baudios (19200, 38400, 57600 o 115200) y el protocolo (Modbus RTU o ASCII) para el puerto serie. Setup base (Configurar base) Seleccione la base del muestreador (refrigerado para todo tipo de clima, refrigerado, estándar portátil o compacto portátil). Language (Idioma) Seleccione el idioma del controlador. Liq sensor cal (Calibración del sensor de líquido)] Calibre el sensor de líquidos o ajústelo a la configuración predeterminada. Set contrast (Ajustar contraste) Ajuste el contraste de la pantalla LCD. Pump duty cycle (Ciclo de trabajo de la bomba) Cambie el ciclo de trabajo de la bomba. Rango: entre 50% y 100% (valor predeterminado: 100%). 68 Español Opción Descripción Password setup (Configurar contraseña) Defina una contraseña para controlar el acceso a los menús de configuración del sistema y del programa. Seleccione ENABLE>YES (Activar>Sí) para establecer una nueva contraseña o ENABLE>NO (Activar>No) para activar la contraseña ya existente. Si se define la contraseña por primera vez, utilice 900900 como la contraseña actual. Tubing life (Vida de la tubería) Defina un recordatorio que indique cuándo se deben cambiar los tubos de la bomba. Desplácese a ENABLE>RESET CYCLES (Activar>Restablecer ciclos) para poner a cero (0) el recuento de ciclos de la bomba. Vaya a ENABLE>CYCLE LIMIT (Activar>Límite de ciclo) para cambiar el número de ciclos de la bomba que debe emplear el recordatorio. Desplácese a ENABLE>STATUS (Activar>Estado) para mostrar el número actual de recuentos de ciclo de la bomba. SDI–12 (optional) [SDI–12 (opcional)] Configure el funcionamiento de un dispositivo SDI–12. Esta opción se muestra únicamente cuando se encuentra un dispositivo. Configuración de un programa de muestreo En un programa de muestreo se incluyen todos los parámetros que son necesarios para recoger muestras de forma automática. Recoja las muestras a intervalos regulares o cuando se complete el programa de muestreo. 1. Pulse MENU (Menú). 2. Desplácese a PROGRAM SETUP>MODIFY ALL (Configuración del programa>Modificar todo). Aparecerá la primera pantalla. 3. Realice una selección o introduzca los valores para cada parámetro. Use las teclas de flecha izquierda y derecha para desplazarse hacia la izquierda o hacia la derecha. Use las teclas de flecha arriba y bajo para cambiar un valor. Si es necesario, cambie las unidades (galones o ml) usando las teclas de flecha. Opción Descripción Bottles (Botellas) Cantidad de botellas: número de botellas que hay en el muestreador (1, 2, 4, 8 o 24). Full BTL shutoff: Enabled/Disabled (Dispositivo de corte para botella llena: activado/desactivado). Si está activado, el controlador comprueba si la botella se desborda. Bottle volume (Volumen de la botella): capacidad volumétrica de cada botella. Rango: 50–65 000 ml (0,01 - 17,17 gal) Intake tubing (Tubos de entrada) Tubing length (Longitud de los tubos): longitud de los tubos de entrada desde el filtro hasta el sensor de líquido. Rango: 100–3000 cm p 3–99 pies. Es necesario indicar una longitud exacta para poder obtener un volumen de la muestra preciso. Intake tube type (Tipo del tubo de entrada): diámetro y material de los tubos de entrada (¼ pulg. de vinilo, 3/8 pulg. de vinilo o 3/8 pulg. de teflón). Program delay (Retardo del programa) Enable/disable (Activar/desactivar): si está activado, el programa de muestreo comienza en un momento específico o tras un número de recuentos determinado. Date and time (fecha y hora): fecha y hora en la que empieza el programa (formato de 24 horas). Counts (Recuentos): número de recuentos o impulsos de un caudalímetro cuando el programa comienza (1 a 9999 recuentos). Si el cadenciado de muestreo se establece posteriormente en base al tiempo, el retardo del programa se desactivará. Español 69 Opción Descripción Ritmo muestra/colecc.muestra Recogida de base tiempo: consulte Recogida de base tiempo en la página 70. Nota: Cuando se utiliza CTVV: • • • • • El sensor de líquido está siempre activado. La distribución de la muestra es compuesta. El modo de ejecución es discontinuo. La primera muestra se toma tras el intervalo. La fuente de flujo es siempre una entrada de 4-20 mA. Por base flujo: consulte Recogida por base flujo en la página 71. Sample distribution (Distribución de la muestra) Deliver samples to all>YES (Entregar muestras a todos>Sí) Se entrega cada muestra a todas las botellas. End after last (Terminar tras última): el programa se detiene después de la última muestra (1–999). Continuous (Continuo): el programa continúa hasta que se detiene de forma manual. Deliver samples to all>NO (Entregar muestras a todos>No) Las muestras se entregan a un subconjunto de botellas . Samples/bottle (Muestras/botellas): número de muestras que hay que recoger en cada botella (1-999). Bottles/sample (Botellas/muestras): número de botellas que contendrán la misma muestra. Liquid sensor (Sensor de líquido) Enable (Activado) o disabled (desactivado). Si está deshabilitado, el volumen de la muestra debe ser calibrado por tiempo. Sample volume (Volumen de la muestra) El volumen de cada muestra en ml (1 galón = 3785,4 ml). Si se está utilizando el modo botellas/muestra, cada botella recibe un volumen completo de la muestra (10-10.000 ml). Asegúrese de que el volumen de la muestra no exceda el volumen de la botella. Los volúmenes de la muestra se redondean a los 10 ml más cercanos. Intake rinses (Lavados del tubo de entrada) El número de veces que el tubo de entrada se lava antes de recoger una muestra (0–3). Sample retries (Reintentos de muestreo) El número de veces que se intenta un muestreo antes de que se produzca un fallo de muestreo (0–3). Site ID (ID del sitio) El nombre de la ubicación de muestreo (máximo 12 caracteres). Si se guarda el programa, la ID del sitio se usa como el nombre preestablecido para el mismo. muestreo avanzado Va al menú de muestreo avanzado. Si es necesario, configure las opciones en el menú Advanced Sampling (Muestreo avanzado). Ejec. programa (Run program) Finaliza la creación del programa de muestreo básico. El tomamuestras pide al usuario que inicie o cancele el programa. Recogida de base tiempo 1. Seleccione o introduzca los valores para cada opción. Opción Descripción Intervalo Recoja las muestras a intervalos de tiempo regulares, en ho ras y minutos (0:01–999:00). 70 Español Opción Descripción Tome 1a muestra El programa se inicia de inmediato o tras el primer intervalo. Volumen variable Especifique si varían los volúmenes de muestras (por caudal). En caso afirmativo, consulte Uso de un volumen variable en la página 71. En caso negativo, consulte Sin uso de un volumen variable en la página 71. Uso de un volumen variable El ritmo de muestra se expresa en tiempo constante volumen variable (CTVV). 1. Seleccione o introduzca los valores para cada opción. Opción Descripción Selecc. unidad Seleccione las unidades de caudal que desea utilizar. Entr. 4mA MAP Introduzca la velocidad de caudal correspondiente para la entrada de 4 mA. Entr. 20mA MAP Introduzca la velocidad de caudal correspondiente para la entrada de 20 mA. Caudal promedio Especifique la velocidad de caudal media de la instalación (1–999 999). Las unidades se especifican en función de las unidades seleccionadas arriba. Time volume desired (volumen y tiempo deseado) Especifique el volumen total que debe recogerse durante el programa de muestreo (10–10 000 ml). Collection period (Período de recogida) Especifique el período de recogida total en horas y minutos (0:01–999:00). Vol min muestreo Si el volumen del caudal no es suficiente para recoger este volumen mínimo, se omitirá la muestra (10–10 000 ml). Sin uso de un volumen variable 1. Seleccione o introduzca los valores para la opción. Opción Descripción Tome 1a muestra Seleccione si desea que el programa se inicie de inmediato o tras el primer intervalo de tiempo. Recogida por base flujo Especifique la fuente de flujo para la toma de muestras de base flujo: • Conteo: consulte Uso de recuentos para determinar la fuente de flujo en la página 71. • Entr. 4-20 mA: consulte Uso de una entrada de 4-20 mA para la fuente de flujo en la página 72. Uso de recuentos para determinar la fuente de flujo 1. Seleccione o introduzca los valores para cada opción. Opción Descripción Muestra cada Recoja muestras a intervalos de caudal regulares, en recuentos (recuentos de 1–9999). Tiempo anulac. Recoja una muestra si el volumen de caudal es muy bajo. Español 71 Opción Descripción Tiempo Tiempo máximo entre muestras (0:01–999:00). El temporizador vuelve a comenzar tras haber recogido cada muestra. Tome 1a muestra El programa se inicia de inmediato o tras el primer intervalo de tiempo. Uso de una entrada de 4-20 mA para la fuente de flujo 1. Seleccione o introduzca los valores para cada opción. Opción Descripción Selecc. unidad Seleccione las unidades de caudal que desea utilizar. Entr. 4mA MAP Introduzca la velocidad de caudal correspondiente para la entrada de 4 mA. Entr. 20mA MAP Introduzca la velocidad de caudal correspondiente para la entrada de 20 mA. Muestra cada Introduzca el caudal total a partir del cual se obtendrá la muestra. Tiempo anulac. Seleccione Habilitar para recoger una muestra si el volumen del caudal es anormalmente bajo. Tiempo Tiempo máximo entre muestras (0:01–999:00). El temporizador vuelve a comenzar tras haber recogido cada muestra. Volumen muestra Ingresar el volumen a recoger por muestra. Si está seleccionado el modo Botella/muestra, cada botella recibe un volumen completo de la muestra (100-10 000 ml). Esta opción se omite si se selecciona un volumen variable. Restauración a los ajustes predeterminados La opción de restauración establece todos los parámetros del programa a los ajustes predeterminados. 1. Pulse MENU (Menú). 2. Desplácese a PROGRAM SETUP>RESTORE (Configuración del programa>Restauración). 3. Seleccione YES (Sí). Iniciar o parar un programa Use la tecla RUN/HALT PROGRAM (Ejecutar/Detener programa) para iniciar o parar un programa. El usuario puede definir hasta 12 horas/fechas de inicio/parada. Para poder modificar la configuración de un sistema o programa hay que detener antes el programa. 1. Pulse RUN/HALT PROGRAM (Ejecutar/Detener programa). 2. Seleccione una de las opciones. Opción Descripción Start (Iniciar) Inicia el programa cargado actualmente. La indicación de estado pasa a RUNNING (En ejecución). Halt (Detener) Detiene temporalmente la ejecución del programa. La indicación de estado pasa a HALTED (Detenido). Resume (Reanudar) El programa reanuda su ejecución desde el punto en el que se detuvo. La indicación de estado pasa a RUNNING (En ejecución). Start over (Reiniciar) El programa comienza desde el principio. La indicación de estado pasa a RUNNING (En ejecución). End program (Terminar el programa) Detiene el programa que se está ejecutando actualmente. La indicación de estado pasa a COMPLETE (Finalizado). 72 Español Operación manual Use la operación manual para recoger una muestra puntual, mover el brazo del distribuidor u operar la bomba. Si desea obtener más información, consulte la versión ampliada de este manual. Calibración del volumen El volumen de la muestra se puede calibrar de forma automática con el sensor de líquido o de forma manual. Cuando el sensor de líquido se usa con el fin de calibrar el volumen, queda ajustado para medir con exactitud todos los volúmenes programados. Si el sensor de líquido está desactivado, el volumen para las muestras, los lavados y las aguas de tormenta se deben calibrar manualmente. En las aplicaciones típicas, se puede usar agua corriente para calibrar el volumen. Si la composición de la muestra difiere significativamente de la composición de las muestras de agua típicas, calibre el tomamuestras usando el líquido que se vaya a recoger. Se deben detener todos los programas antes de efectuar la calibración. Calibración del sensor de líquido La precisión del sensor de líquido varía según el tipo de líquido que se esté muestreando. Por ejemplo, el volumen de una muestra muy turbia puede no ser tan exacto como el del agua limpia. El sensor de líquido se puede calibrar con el agua de muestreo para mejorar la precisión del volumen. 1. Coloque los tubos de admisión dentro del agua de muestreo. 2. Pulse MENU (Menú). 3. Para comprobar que el sensor de líquido está activado desplácese a PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuración del programa>Modificar selección>Sensor de líquido>Activar). 4. Vaya a SYSTEM SETUP>LIQ SENSOR CAL (Configuración del sistema>Calibración del sensor de líquido). 5. Seleccione PERFORM CAL (Efectuar calibración). 6. Seleccione START (Iniciar). La bomba funcionará a la inversa para purgar el tubo de admisión. A continuación, la bomba funcionará en dirección hacia delante. 7. Observe el caudal de muestra del tubo de admisión para comprobar que la muestra pasa por el sensor de líquido. 8. Cuando la muestra pase a través del sensor de líquido y no haya burbujas, pulse la tecla STOP (Detener). El tubo de admisión se purga automáticamente. La calibración del sensor de líquido se ha efectuado. 9. Seleccione DONE (Listo) para salir o REPEAT (Repetir) para volver a realizar la calibración. Restauración a los valores de calibración predeterminados Es posible hacer que la calibración del volumen del sensor de líquido vuelva a tener los ajustes predeterminados de fábrica. 1. Para comprobar que el sensor de líquido está activado desplácese a PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuración del programa>Modificar selección>Sensor de líquido>Habilitar). 2. Pulse VOLUME CALIBRATION (Calibración del volumen) y seleccione RESET CAL (Restablecer calibración). La calibración del volumen está especificada según los ajustes de fábrica. Aparecerá el mensaje “value has been reset to 0” (el valor se ha restablecido a cero). Calibración del volumen de la muestra con el sensor de líquido Calibre el volumen de la muestra con el sensor de líquido activado para ajustar levemente los volúmenes de muestra. Español 73 1. Para comprobar que el sensor de líquido está activado desplácese a PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuración del programa>Modificar selección>Sensor de líquido>Activar). 2. Introduzca los tubos de admisión en el agua de muestro o en agua corriente. Nota: Para obtener una mejor precisión, use el agua de muestreo. 3. Quite los tubos del lado de salida de la bomba e introdúzcalos en una probeta. 4. Pulse VOLUME CALIBRATION (Calibración del volumen) y seleccione CALIBRATION (Calibración). 5. Seleccione START (Iniciar). La bomba funcionará a la inversa para purgar el tubo de admisión. A continuación, la bomba funcionará en dirección hacia delante y recogerá el volumen de la muestra especificado en el programa de muestreo. La bomba funcionará a la inversa para purgar el tubo de admisión. 6. Cuando se haya recogido la muestra, seleccione DONE (Listo). Compare el volumen de la muestra de la probeta con el volumen de la muestra introducido en el menú de configuración del programa. 7. Si el volumen que fue recogido difiere del volumen de muestra indicado en el programa actual, especifique el volumen que fue recogido realmente. Pulse ENTER (Intro). El sensor se ajusta para medir con precisión todos los volúmenes programados. 8. Seleccione START (Iniciar) para comenzar el programa de muestreo o pulse CANCEL (Cancelar) para salir del menú de calibración del volumen. Calibración del volumen de la muestra de forma manual Si el sensor de líquido está desactivado, hay que calibrar todos los volúmenes de forma manual y por tiempo. El volumen de muestra se calibra para el volumen especificado en el programa actual. Si se cambia el volumen de la muestra en el programa, habrá que volver a calibrar el volumen de la muestra para el nuevo valor. 1. Para comprobar que el sensor de líquido está desactivado desplácese a PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuración del programa>Modificar selección>Sensor de líquido>Desactivar). 2. Introduzca los tubos de admisión en el agua de muestro o en agua corriente. 3. Quite los tubos del lado de salida de la bomba e introdúzcalos en una probeta. 4. Pulse VOLUME CALIBRATION (Calibración del volumen). Si aparece más de un volumen, seleccione uno de los volúmenes para realizar la calibración. 5. Seleccione START (Iniciar). La bomba funcionará a la inversa para purgar el tubo de admisión. A continuación, la bomba funcionará en dirección hacia delante y comenzará a recoger la muestra. 6. Pare la bomba en el volumen o lugar de lavado seleccionado: • Volumen de la muestra: pulse STOP (Detener) cuando se haya recogido el volumen que se especifica en el programa principal. • Lavado: pulse STOP (Detener) cuando el líquido de los tubos de admisión llegue al detector de líquido. • Volumen de tormenta: pulse STOP (Detener) cuando se haya recogido el volumen que se especifica en el programa de aguas de tormenta. 7. Seleccione DONE (Listo) para pasar a la siguiente calibración del volumen o pulse REPEAT (Repetir) para volver a efectuar la calibración. 8. Cuando se hayan calibrado todos los volúmenes, pulse DONE (Listo). 9. Conecte los tubos de salida a la conexión del tubo del tomamuestras. 10. Seleccione START (Iniciar) para comenzar el programa de muestreo o CANCEL (Cancelar) para salir. 74 Español Verificación del volumen de la muestra Para comprobar la precisión del volumen de la muestra, tome una muestra al azar. No vuelva a la calibración para verificar el volumen, porque la compensación del volumen se restablece a 0 al inicio de la calibración. 1. 2. 3. 4. 5. 6. 7. 8. Pulse MANUAL OPERATION (Funcionamiento manual). Seleccione GRAB SAMPLE (Muestra al azar). Coloque los tubos de admisión dentro del agua de muestreo. Quite los tubos del lado de salida de la bomba e introdúzcalos en una probeta. Si el sensor de líquido está activado, introduzca el volumen que va a verificar. Si el sensor de líquido está desactivado, introduzca el volumen que se especifica en el programa. Pulse ENTER (Intro). Se inicia el ciclo de bomba (purga-muestra-purga). Compare el volumen que se ha recogido en la probeta con el valor del volumen que se esperaba. Si el volumen recogido no es el correcto, vuelva a efectuar la calibración del volumen. Mantenimiento PELIGRO Peligros diversos. Sólo el personal cualificado debe realizar las tareas descritas en esta sección del documento. PELIGRO Peligro de electrocución. Retire la alimentación eléctrica del instrumento antes de realizar actividades de mantenimiento o reparación. ADVERTENCIA Exposición a riesgo biológico. Siga protocolos de manipuleo seguros mientras esté en contacto con las botellas y los componentes del tomamuestras. ADVERTENCIA Peligros diversos. El técnico debe asegurarse de que el equipo funciona de forma segura y correcta después de los procedimientos de mantenimiento. Limpieza del instrumento PRECAUCIÓN Peligro de incendio. No utilice agentes inflamables para limpiar el instrumento. AVISO No limpie el calefactor del compartimiento del controlador con ningún tipo de líquido. Si el controlador y la bomba no se limpian bien solo con agua, desconecte el controlador y aléjelo del muestreador. Espere el tiempo suficiente a que el controlador y la bomba se sequen antes de volver a instalar las piezas y ponerlas en funcionamiento de nuevo. Limpie el tomamuestras de la siguiente manera: • Carcasa del muestreador—Limpie la superficie interna y externa con un paño húmedo y detergente suave. No utilice limpiadores abrasivos ni disolventes. Español 75 • Botellas para muestras—Limpie las botellas con un cepillo, agua y un detergente suave. Enjuáguelas con agua corriente y, a continuación, con agua destilada. Si es necesario, utilice un autoclave para limpiar las botellas de vidrio. Cambio del desecante Un cartucho desecante en el controlador absorbe la humedad y evita la corrosión. Controle el color del desecante a través de la ventana (Figura 12). El desecante en buen estado es de color amarillo. Cuando el color sea verde, cambie el desecante. 1. 2. 3. 4. 5. 6. Desatornille y quite el cartucho de desecante (Figura 12). Quite el tapón y tire el desecante gastado. Llene dicho tubo con desecante nuevo. Coloque el tapón. Aplique grasa a la junta tórica. Coloque el tubo de desecante en el controlador. Figura 12 Cartucho desecante 1 Tapa 3 Junta tórica 2 Tubo con desecante 4 Ventana para observar el desecante Mantenimiento de la bomba PRECAUCIÓN Riesgo de opresión. Retire la alimentación eléctrica del instrumento antes de realizar actividades de mantenimiento o reparación. Sustitución de los tubos de la bomba AVISO Si se utilizan tipos distintos a los que suministra el fabricante puede provocarse un desgaste excesivo en las piezas mecánicas o un mal rendimiento de la bomba. Compruebe si los tubos de la bomba están desgastados en la zona donde los rodillos rozan con los tubos. Sustituya los tubos cuando aparezca el mensaje de recordatorio o cuando estos muestren indicios de desgaste. Requisitos previos: • Tubos de la bomba: precortados o a granel, 4,6 m o 15,2 m (15 pies o 50 pies) 1. Desconecte la alimentación del controlador. 2. Si se utilizan tubos a granel, córtelos y agregue los puntos de alineación. Consulte el manual completo en la página web del fabricante. 3. Quite los tubos de la bomba, como se muestra en los siguientes pasos ilustrados. 76 Español 4. Limpie los residuos de silicona del interior del alojamiento de la bomba y de los rodillos. 5. Coloque los nuevos tubos de la bomba, como se muestra en los siguientes pasos ilustrados. Español 77 Limpieza del rotor Limpie el rotor, los encauzadores del tubo de la bomba y el alojamiento de la bomba con un detergente suave. Consulte el apartado Sustitución de los tubos de la bomba en la página 76 y los siguientes pasos ilustrados. Sustitución del tubo del brazo del distribuidor El brazo del distribuidor se desplaza por encima de cada botella durante el muestreo con múltiples botellas. Sustituya el tubo del brazo del distribuidor cuando el tubo esté desgastado. Asegúrese de usar el tubo correcto para el distribuidor y el brazo del distribuidor correctos. Nota: El tubo del distribuidor es diferente del tubo de la bomba. Si se instala el tubo de la bomba en el conjunto del distribuidor, puede dañar el distribuidor. Además, se pueden perder muestras porque resulta difícil mover el brazo del distribuidor. 1. Retire el tubo del brazo del distribuidor y del techo de la sección central. 2. Introduzca el nuevo tubo en el brazo del distribuidor. Asegúrese de que el tubo está alineado conforme a las medidas que se muestran en la Figura 13. 3. Introduzca el otro extremo del tubo en la conexión del techo de la sección central. 4. Lleve a cabo una prueba de diagnóstico del distribuidor para comprobar que el funcionamiento es correcto. Figura 13 Conjunto distribuidor 1 Boquilla 3 Motor del distribuidor 2 Brazo del distribuidor 4 Eje 78 Español 5 Tubería del distribuidor Sustitución de fusibles El fusible no es una pieza que pueda reparar el usuario. Póngase en contacto con el servicio de asistencia técnica. Un fusible fundido podría indicar que hay un problema en el muestreador y que se necesita mantenimiento. Español 79 Índice Especificações na página 80 Operação na página 94 Informações gerais na página 82 Manutenção na página 100 Instalação na página 85 Informações adicionais Você poderá encontrar informações adicionais no website do fabricante. Especificações As especificações estão sujeitas a alterações sem aviso prévio. Amostrador Portátil SD900 Especificação Base padrão Base compacta Base de composto Dimensões Diâmetro: 50,5 cm (19,9 pol.) Diâmetro: 44,1 cm (17,4 pol.) Diâmetro: 50,28 cm (19,8 pol.) Altura: 61 cm (24 pol.) Altura: 79,5 cm (31,4 pol.) 12,2 kg (27 lb) com garrafas de polietileno de 575 mL (19,44 oz) (24x) 15 kg (36 lb) com garrafas de polietileno de 950 mL (32,12 oz) (12x) Altura: 69,4 cm (27,3 pol.) Peso 15 kg (35,6 lb) com garrafas de polietileno (24x) 14,8 kg (32,6 lb) com recipiente de polietileno de 10 L (2,5 gal) (1x) 12,9 kg (28,3 lb) com recipiente de polietileno de 10 L (2,5 gal) (1x) Caixa ABS resistente a impacto, construção de 3 seções, base com paredes duplas com 2,54 cm (1 pol.) isolamento — contato direto da garrafa com gelo. Temperatura da amostra 0 a 60 °C (32 a 140 °F) Filtros Aço inoxidável 316 em tamanho padrão, alta velocidade ou perfil baixo para aplicativos de profundidade rasa e Teflon® ou aço inoxidável 316 em tamanho padrão Tubulação de entrada de amostra 9,5 mm (3/8 pol.) Polietileno de vinil I.D. ou revestido de Teflon Capacidade da garrafa de amostra Garrafas de polietileno de 1 L (0,26 gal) e/ou de vidro de 350 mL (11,83 oz) (24x) Garrafas de polietileno de 2,3 L (0,6 gal.) e/ou de vidro de 1,9 L (0,5 gal.) (8x) Garrafas de polietileno de 3,8 L (1 gal.) e/ou de vidro de 3,8 L (1 gal.) (4x) Garrafas de polietileno de 3,8 L (1 gal.) e/ou de vidro de 3,8 L (1 gal.) (2x) Recipiente composto de polietileno de 21 L (5,5 gal) ou 15 L (4 gal) ou polietileno de 20 L (5,25 gal) ou polietileno de 10 L (2,5 gal) ou vidro de 10 L (2,5 gal) (1x) 80 Português Garrafas de polietileno de 575 mL (19,44 oz) (24x) Garrafas de vidro de 950 mL (32,12 oz) (8x) Garrafa de polietileno de 10 L (2,5 gal) (1x) Garrafa de vidro de 10 L (2,5 gal) (1x) Garrafa de polietileno de 21 L (5,5 gal) (1x) controlador SD Especificação Detalhes Dimensões (L x A x P) 29,2 x 17,1 x 26.4 cm (11½ x 6¾ x 103/8 pol.) Peso 4.2 kg (9.26 lb) Caixa Mescla PC/ABS, NEMA 4X, 6, IP 67 Alimentação elétrica 15 Vcd alimentado por uma fonte de alimentação 8754500; 15 Vcd alimentado por uma fonte de alimentação integral Proteção de sobrecarga Fusível de linha CD de 7 A para a bomba Temperatura de operação 0 a 50°C (32 a 122°F) Temperatura de armazenamento –30 a 60°C (-22 a 140°F) Umidade de armazenamento/operação Condensação de 100% Bomba Alta velocidade peristáltica, com cilindros de Nylatron montados em molas Caixa da bomba IP37 Tubo da bomba Silicone com DI de 9,5 mm x DE de 15,9 OD mm (3/8 pol. de DI x 5/8 pol. de DE) Vida da tubulação da bomba 20.000 ciclos de amostra com: volume de amostra de 1 l (0,3 gal.), 1 lavagem, intervalo com ritmo de 6 minutos, 4,9 m (16 pés) de tubo de entrada de 3/8 pol., 4,6 m (15 pés) de elevação vertical, temperatura de amostra de 21 °C (70 °F) Elevação vertical de amostra Máximo 8,5 m (28 pés) para: 8,8 m (29 pés) de tubo de entrada de vinil de 3/8 pol. ao nível do mar a 20–25 °C (68–77 °F) Taxa de fluxo da bomba 4,8 l/min (1,25 gpm) na elevação vertical de 1 m (3 pés) com tubo de entrada de 3/8 pol. típico Volume de amostra Programável em incrementos de 10 ml (0,34 oz) de 10 a 10.000 ml (3,38 oz. a 2,6 gal.) Repetibilidade de volume de amostra (típica) ±5% de 200 ml volume da amostra usando detecção de líquido sem calibragem com: com suspensão vertical de 4,6 m (15 pés), 4,9 m (16 pés) de tubo de entrada de vinil de 3/8 pol., garrafa única, válvula de garrafa cheia em temperatura ambiente e elevação de 1524 m (5000 pés) Precisão de volume de amostra (típica) ±10% de volume da amostra de 200 ml usando detecção de líquido sem calibragem com: com suspensão vertical de 4,6 m (15 pés), 4,9 m (16 pés) de tubo de entrada de vinil de 3/8 pol. garrafa única, válvula de garrafa cheia em temperatura ambiente e elevação de 1524 m (5000 pés) Modos de amostragem Passo: fixo-tempo, fixo-fluxo, variável-tempo, variável-fluxo, volume variável de tempo constante (CTVV). Consulte a amostragem Volume variável de tempo constante (CTVV) na versão estendida deste manual. Distribuição: composto de frasco único, composto de vários frascos, separado de vários frascos, frascos por amostra, amostras por frasco, combinação de frascos por amostra, amostras por frasco Modos de execução Contínuo ou não-contínuo com número de amostras inserido pelo usuário Programas múltiplos Armazena até três programas de amostragem Velocidade de transferência (típica) 0,9 m/s (2,9 pés/s) com: elevação vertical de 4,6 m (15 pés), 4,9 m (16 pés) de tubulação de entrada de vinil de 3/8 pol. 21 °C (70 °F) e elevação de 1524 m (5000 pés) Português 81 Especificação Detalhes sensor de líquido Ultrasônico. Corpo: Ultem NSF ANSI aprovado pelo padrão 51, de acordo com a USP Classe VI histórico de amostra Até 510 registros Purga de ar Ar purgado automaticamente antes e depois de cada amostra. A duração compensa automaticamente as variações no comprimento dos tubos de entrada. Registro de eventos 510 registros ® Conexões Materiais molhados Aço inoxidável, polietileno, Teflon, Ultem, silicone Comunicações RS232, Modbus, SDI-12 Garantia 1 ano Informações gerais Em hipótese alguma o fabricante será responsável por danos diretos, indiretos, especiais, incidentais ou consequenciais resultantes de qualquer defeito ou omissão neste manual. O fabricante reserva-se o direito de fazer alterações neste manual e nos produtos aqui descritos a qualquer momento, sem aviso ou obrigação. As edições revisadas podem ser encontradas no site do fabricante. Informações de segurança AVISO O fabricante não é responsável por quaisquer danos devido ao uso ou aplicação incorreta deste produto, incluindo, sem limitação, danos diretos, acidentais ou consequenciais, e se isenta desses danos à extensão total permitida pela lei aplicável. O usuário é unicamente responsável por identificar riscos críticos de aplicação e por instalar os mecanismos apropriados para proteger os processos durante um possível mau funcionamento do equipamento. Leia todo o manual antes de tirar da embalagem, montar ou operar esse equipamento. Preste atenção a todas as declarações de perigo e cuidado. Caso contrário, o operador poderá sofrer ferimentos graves ou o equipamento poderá ser danificado. Certifique-se de que a proteção oferecida por este equipamento não seja afetada. Não use nem instale este equipamento de nenhuma outra forma além da especificada neste manual. Uso de informações de risco PERIGO Indica uma situação potencial ou iminentemente perigosa que, se não for evitada, resultará em morte ou lesão grave. ADVERTÊNCIA Indica uma situação potencialmente perigosa que, se não for evitada, pode resultar em morte ou ferimento grave. CUIDADO Indica uma situação potencialmente perigosa que pode resultar em ferimento leve a moderado. AVISO Indica uma situação que, se não evitada, pode causar danos ao instrumento. Informações que necessitam de uma ênfase especial. 82 Português Avisos de precaução Leia todas as etiquetas e rótulos fixados no instrumento. Caso não sejam observadas, podem ocorrer lesões pessoais ou danos ao instrumento. Um símbolo no instrumento tem sua referência no manual com uma medida preventiva. Este é o símbolo de alerta de segurança. Acate todas as mensagens de segurança que seguem este símbolo a fim de evitar lesões potenciais. Se o símbolo estiver no instrumento, consulte o manual de instruções para obter informações sobre a operação ou segurança. Este símbolo indica possível risco de pinçamento. O equipamento elétrico marcado com este símbolo não pode ser descartado em sistemas de descarte público ou doméstico europeus. Devolva equipamentos antigos ou no final da vida útil para o fabricante para descarte, sem custo adicional para o usuário. Certificação Canadian Radio Interference-Causing Equipment Regulation (Regulamentação para equipamentos de rádio causadores de interferência do Canadá), IECS-003, Classe A: Os registros de testes de comprovação encontram-se com o fabricante. Este aparelho digital Classe A atende a todos os requisitos de regulamentações canadenses sobre equipamentos que causam interferências. Cet appareil numèrique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences. FCC parte 15, limites Classe "A" Os registros de testes de comprovação encontram-se com o fabricante. O dispositivo está em conformidade com a Parte 15 das Regras da FCC. A operação está sujeita às seguintes condições: 1. O equipamento não deve causar interferência prejudicial. 2. O equipamento deve aceitar todas as interferências recebidas, inclusive interferências que podem causar funcionamento indesejado. Alterações ou modificações a este equipamento não aprovadas expressamente pela parte responsável pela conformidade podem anular a autoridade do usuário de operar o equipamento. Este equipamento foi testado e está em conformidade com os limites de dispositivo digital Classe A, de acordo com a Parte 15 das Regras da FCC. Esses limites foram estabelecidos para proporcionar uma razoável proteção contra interferências nocivas quando o equipamento for operado em ambientes comerciais. Este equipamento gera, utiliza e pode irradiar energia de rádiofrequência e, se não instalado e usado de acordo com o manual de instruções, pode causar interferências prejudiciais às comunicações de rádio. É provável que o funcionamento deste equipamento em área residencial possa causar interferência indesejada, caso em que o usuário será solicitado a corrigir a interferência por conta própria. As seguintes técnicas podem ser usadas para reduzir problemas de interferência: 1. Desconecte o equipamento de sua fonte de alimentação para verificar se ele é ou não a origem da interferência. 2. Se o equipamento está conectado à mesma tomada do dispositivo que está sofrendo interferência, conecte o equipamento a uma tomada diferente. 3. Afaste o equipamento do dispositivo que estiver recebendo a interferência. 4. Reposicione a antena de recebimento do dispositivo que está sofrendo interferência. 5. Tente algumas combinações das opções acima. Português 83 Visão geral do produto PERIGO Riscos químicos ou biológicos. Se esse instrumento for usado para monitorar um processo de tratamento e/ou sistema de alimentação química para o qual existam limites de regulamentação e requisitos de monitoramento relacionados à saúde pública, à produção ou ao processamento de alimentos ou bebidas, é responsabilidade do usuário deste instrumento conhecer e cumprir as regulamentações aplicáveis e ter mecanismos suficientes e apropriados para obter conformidade com as regulamentações aplicáveis no caso de mau funcionamento do instrumento. CUIDADO Perigo de incêndio. Este produto não foi projetado para uso com líquidos inflamáveis. O Amostrador Portátil SD900 coleta amostras de líquidos em intervalos específicos e mantém as amostras em garrafas ou recipientes. Use o amostrador para uma variedade de aplicações aquosas e para poluentes tóxicos e sólidos suspensos. Configure o amostrador com diferentes retentores, garrafas e recipientes. Consulte Figura 1. Figura 1 Visão geral do produto 1 Base compacta 6 Conexões do controlador 2 Base isolada padrão 7 Bomba 3 Seção central 8 Controlador 4 Fonte de energia 9 sensor de líquido 5 Cobertura superior 10 Recipiente de base isolada padrão para 21 L (5,5 gal) Componentes do produto Certifique-se de que todos os componentes foram recebidos. Consulte Figura 2. Se houver itens ausentes ou danificados, entre em contato imediatamente com o fabricante ou com um representante de vendas. 84 Português Figura 2 Componentes do produto 1 Base (padrão, compacta ou composta) 6 Carregador de bateria (opcional) 2 Componentes para uma opção de garrafa única (garrafa e suporte podem mudar) 7 Suprimento de força CA (opcional) 3 Seção central com controlador 8 Filtro 4 Bateria (opcional) 9 Tubulação de entrada, coberta de vinil ou teflon 5 Cobertura superior 10 Componentes para uma opção de múltiplas garrafas (as garrafas e os retentores podem mudar) Instalação PERIGO Vários perigos. Somente pessoal qualificado deve realizar as tarefas descritas nesta seção do manual. Precauções em espaços confinados PERIGO Perigo de explosão. Treinamento em testes pré-entrada, ventilação, procedimentos de entrada, procedimentos de evacuação/resgate e práticas de trabalho de segurança são necessárias antes de entrar em espaços confinados. As informações a seguir são fornecidas para ajudar os usuários a entenderem os perigos e os riscos associados com a entrada em espaços confinados. Em 15 de abril de 1993, a decisão final da OSHA sobre o CFR 1910.146, Autorização Requerida para Espaços Confinados, se tornou lei. Este padrão afeta diretamente mais de 250.000 locais industriais nos EUA e foi criado para proteger a saúde e a segurança dos trabalhadores em espaços confinados. Definição de um espaço confinado: Um espaço confinado é qualquer local ou recinto que apresente (ou tenha potencial imediato para apresentar) uma ou mais das seguintes condições: • Uma atmosfera com uma concentração de oxigênio menor que 19,5% ou maior que 23,5% e/ou uma concentração de sulfeto de hidrogênio (H2S) que seja maior que 10 ppm. • Uma atmosfera que possa ser inflamável ou explosiva devido a gases, vapores, névoas, poeira ou fibras. Português 85 • Materiais tóxicos que, mediante contato ou inalação, podem causar lesões, danos à saúde ou morte. Os espaços confinados não são feitos para ocupação humana. Os espaços confinados têm uma entrada restrita e contêm riscos conhecidos ou potenciais. Exemplos de espaços confinados incluem câmaras subterrâneas, chaminés, tanques, subterrâneos de troca e outros locais semelhantes. Os procedimentos de segurança padrão devem sempre ser obedecidos antes da entrada nos espaços confinados e/ou locais onde possam estar presentes gases perigosos, vapores, névoas, poeiras ou fibras. Antes de entrar em um local confinado, encontre e leia todos os procedimentos relacionados à entrada em um espaço confinado. Instalação mecânica Diretrizes da instalação no local PERIGO Risco de explosão. O instrumento não está aprovado para ser instalado em localizações perigosas. Consulte as diretrizes que se seguem para a avaliação da localização do site. • Obedeça a todas as precauções de segurança se o amostrador for instalado em um local confinado. Consulte Precauções em espaços confinados na página 85. • Certifique-se de que a temperatura no local está no intervalo específico. Consulte Especificações na página 80. • Instale o amostrador em uma superfície nivelada ou pendure o amostrador com o suporte de suspensão, a alça de suporte ou a barra de abertura. Consulte Instale o amostrador em um bueiro na página 86 e a documentação da instalação aplicável. • Esteja o mais próximo possível da origem da amostra para reduzir a demora da análise. Consulte Instale a tubulação do amostrador na página 89. • Para obter as limitações sobre a velocidade do transporte e máxima elevação vertical, consulte Especificações na página 80. Instale o amostrador em um bueiro Instale o amostrador acima da água de amostra em um bueiro. Instale o amostrador com uma barra de abertura ou uma alça de suporte. Instale a barra de abertura dentro do bueiro. A barra de abertura é suportada pela pressão contra as paredes. A alça de suporte tem a mesma largura que a tampa do bueiro. Instale a alça de suporte diretamente abaixo da tampa para obter suporte.Consulte o manual expandido no site do fabricante.Consulte a documentação fornecida com os acessórios para instalar o amostrador. Instale o distribuidor ou o fechamento total da garrafa (opcional) O conjunto do distribuidor ou o fechamento total da garrafa é normalmente instalado na fábrica. Consulte a documentação do distribuidor ou do fechamento total da garrafa para a instalação. Prepare o amostrador Limpe as garrafas de amostra Limpe as garrafas de amostra e as tampas com escova, água e detergente suave. Lave os recipientes com água corrente e, em seguida, enxágue com água destilada. Instale uma única garrafa Use uma única garrafa para coletar uma amostra composta. Quando a garrafa estiver cheia, a válvula de garrafa cheia para o programa de amostra. 1. Limpe as garrafas de amostra. Consulte Limpe as garrafas de amostra na página 86. 2. Instale a garrafa de amostra como mostrado na Figura 3. 86 Português Figura 3 Instalação de garrafa única 1 Garrafa de polietileno, 10 L (2,5 gal) (1918) 6 Garrafa de polietileno, 21 L (5,5 gal) (6494) 2 Garrafa de vidro, 10 L (2,5 gal) (6559) 7 Base compacta (8975) 3 Suporte (1502) 8 Base isolada padrão (8976) 4 Garrafa de polietileno, 15 L (4 gal) (1367) 9 Garrafa de base isolada padrão para 21 L (5,5 gal) (8561) 5 Garrafa de polietileno, 19 L (5 gal) (6498) Instale a primeira garrafa para várias coleções de amostras Use garrafas múltiplas para coletar amostras em garrafas separadas ou em mais do que uma garrafa. O distribuidor move o tubo de amostra acima de cada garrafa. Instale as garrafas na base do amostrador como indicado na Figura 4. Instale a primeira garrafa de amostra (número 1) abaixo da etiqueta na base do amostrador. Instale as demais garrafas em números crescentes na direção indicada pela etiqueta. Consulte Figura 5 na página 89 para obter um diagrama dos componentes necessários. Português 87 Figura 4 Instalação da garrafa número 1 1 Localização da garrafa número 1 (base compacta) 6 Faixas elásticas 2 Localização da garrafa número 1 para 24 garrafas 7 Garrafa de base isolada padrão para 21 L (5,5 gal) 3 Localização da garrafa número 1 para 8 garrafas 8 Base isolada padrão 4 Localização da garrafa número 1 para 2 ou 4 garrafas 9 Base compacta 5 Retentor Instale várias garrafas Quando várias garrafas estão instaladas, um braço distribuidor move o tubo de amostra sobre cada garrafa. A coleta de amostra é automaticamente parada quando o número especificado de amostras tiver sido coletado. 1. Limpe as garrafas de amostra. Consulte Limpe as garrafas de amostra na página 86. 2. Instale a primeira garrafa de amostra (número 1) abaixo da etiqueta na base do amostrador. Consulte Instale a primeira garrafa para várias coleções de amostras na página 87. 3. Ordene as garrafas de amostra como mostrado na Figura 5. Para oito garrafas ou mais, certifique-se de que colocar a primeira garrafa próxima ao indicador da garrafa um no sentido horário. 4. Coloque o conjunto de garrafas no amostrador. Para oito garrafas ou mais, alinhe a grade nas ranhuras da bandeja de garrafas. 88 Português Figura 5 Instalação de garrafas múltiplas 1 Posicionador/retentor (2347) 8 Garrafas de vidro, 3,8 L (1 gal), (4x) (2216) ou garrafas de polietileno, 3,8 L (1 gal), (4x) (2217) 2 Retentor (1422) 9 Garrafas de vidro, 1,9 L (0,5 gal), (8x) (1118) ou garrafas de polietileno, 2,3 L (0,61 gal), (8x) (657) 3 Retentor (2190) 10 Garrafas de polietileno, 1 L (0,26 gal), (24x) (737) 4 Retentor (2189) 11 Garrafas de vidro, 350 mL (11,83 oz) (24x) (732) 5 Garrafas de vidro, 950 mL (32,12 oz) (8x) (2348) 12 Base compacta (8975) 6 Garrafas de polietileno, 575 mL (19,44 oz), (24x) (1369) 13 Base padrão (8976) 7 Garrafas de vidro, 3,8 L (1 gal), (2x) (2214) ou garrafas de polietileno, 3,8 L (1 gal), (2x) (2215) Instale a tubulação do amostrador Instale o filtro no meio do fluxo da amostra (não próximo à superfície ou ao fundo) para se certificar de que uma amostra representativa seja coletada. Consulte Figura 6. Consulte Figura 7 para a instalação do tubo de entrada. 1. Conecte a tubulação ao amostrador como mostrado na Figura 7. Observação: Use o kit 2186 de conexão se a tubulação revestida de Teflon for usada. 2. Instale o tubo de entrada e o filtro no fluxo principal da fonte de amostra onde a água está turbulenta e bem misturada. • Deixe o tubo de entrada o mais curto possível. • Mantenha o tubo de entrada em uma inclinação mais vertical possível para que seque completamente entre uma amostra e outra. Observação: Se não for possível manter uma inclinação vertical ou se o tubo for pressurizado, desative o sensor de líquido. Calibre o volume de amostra manualmente. Consulte Calibre o volume de amostra manualmente na página 100. • Certifique-se de que o tubo de entrada não esteja comprimido. Português 89 Figura 6 Instalação de instrumento 1 Filtro 3 Tubo de entrada 2 Elevação vertical Figura 7 Instalação do tubo de entrada Instalação elétrica Conexões do controlador ADVERTÊNCIA Risco de choque elétrico. Equipamento conectado externamente deve ter uma avaliação apropriada do padrão de segurança do país. A Figura 8 mostra as conexões que podem ser feitas ao controlador. 90 Português Figura 8 Conectores do controlador 1 Dispositivo auxiliar 4 Opção de dispositivo SDI-12 2 Fonte de alimentação 5 Distribuidor/válvula de garrafa cheia 3 Comunicações seriais Conecte um medidor de fluxo Conecte um medidor de fluxo ao controlador para iniciar ou parar o amostrador quando o fluxo de amostra ficar acima ou abaixo de um valor especificado. Itens para coletar: • Cabo pleno multi-fim para medidores de fluxo Sigma (ou meio cabo 980 para o modelo de medidor de fluxo 980) • Divisor opcional para conexões extras. É possível conectar dois ou mais divisores em série. 1. Conecte uma extremidade do cabo ao medidor de fluxo. Para o medidor de fluxo modelo 980, consulte o manual do usuário do medidor de fluxo modelo 980. 2. Conecte a outra extremidade do cabo ao conector auxiliar do dispositivo no controlador. Observação: Se o medidor de fluxo possuir cabo de 6 pinos, use o cabo adaptador de 6 a 7 pinos. Conecte um PC ou rede de comunicações Conecte um PC ou uma rede de comunicações ao controlador para transferir dados ou para configurar o amostrador. Itens para coletar: • Cabo serial, RS232 de 7 pinos para DB-9 (consulte a Figura 9). Conexões: B no 5 (sinal terra); D no 3 (RCD); F no 2 (TXD); G (terra). 1. Conecte uma extremidade do cabo serial ao dispositivo ou rede de comunicação. 2. Conecte a outra extremidade do cabo ao conector serial de comunicações no controlador. Figura 9 Cabo serial de comunicações Português 91 Conecte o amostrador à alimentação PERIGO Risco de choque elétrico. Se este equipamento for usado ao ar livre ou em locais potencialmente úmidos, um dispositivo contra Falhas de Aterramento (GFCI/GFI, Ground Fault Circuit Interrupt) deve ser usado para conectar o equipamento à sua fonte principal de energia. PERIGO Perigo de incêndio. Instale um disjuntor de 15 A na linha de alimentação. Um disjuntor pode ser o interruptor local, caso esteja bem próximo ao equipamento. PERIGO Risco de eletrocussão. A conexão com aterramento protetor (PE) é obrigatória ADVERTÊNCIA Risco de eletrocussão. Certifique-se de que haja um fácil acesso para desconectar a energia local. ADVERTÊNCIA Risco de choque elétrico. A fonte de alimentação pode se sobreaquecer se o tempo entre os ciclos de bombeamento for muito curto. AVISO Certifique-se de que o período de ciclo da bomba não sobreaquecerá a fonte de alimentação. Consulte Calcule o tempo de ciclo da bomba na página 93. Conecte o amostrador à bateria ou a uma fonte de alimentação AC. Consulte as etapas ilustradas seguintes. 92 Português Calcule o tempo de ciclo da bomba A bomba deve permanecer desligada por um período mínimo de tempo entre os ciclos de amostra, para que a fonte de alimentação não sobreaqueça. 1. Calcule o tempo total de funcionamento contínuo da bomba durante um ciclo de amostra. Inclua todos os estágios: pré-purgação, lavagem de entrada, amostra, tentativas de amostra e póspurgação. 2. Encontre o menor período de tempo em que a bomba deve ficar desligada entre ciclos de amostra na Figura 10. 3. Certifique-se de que o ritmo ou intervalo de tempo do programa amostrador permita que a bomba fique desligada entre ciclos de amostra pela quantidade mínima de tempo. Figura 10 Períodos em que bomba fica ligada e desligada a 50 °C (122 ºF) 1 Tempo (em minutos) em que a bomba deve permanecer desligada 2 Tempo (em minutos) em que a bomba deve permanecer ligada continuamente 3 Exemplo: se a bomba permanecer ligada continuamente por 5 minutos, ela deve ficar desligada por 15 minutos. Ligue a alimentação Pressione o botão de ligação para ligar ou desligar. Certifique-se de que a fonte de alimentação (alimentação AC ou alimentação da bateria) esteja corretamente instalada. Interface do usuário e navegação O teclado do controlador é mostrado na Figura 11. Use a tecla MENU para configurar os programas de amostragem, para definir as configurações do controlador ou para realizar testes de diagnóstico. Use as teclas de seta, ENTER e BACK (Voltar) para rolar pelo menu, fazer seleções e inserir valores. Procure setas na parte inferior ou lateral da tela para saber quando mais opções estão disponíveis. Português 93 Figura 11 Teclado do controlador 1 POWER (Força) 5 MENU 9 STATUS 2 VOLUME CALIBRATION (Calibração de volume) 6 ENTER 10 RUN/HALT PROGRAM (Executar/interromper programa) 3 STOP (Parar) 7 VOLTAR 4 LED 8 TECLAS DE SETAS 11 MANUAL OPERATION (Operação manual) Operação Defina as configurações do sistema 1. Pressione MENU. 2. Vá até SYSTEM SETUP (Configuração do sistema). 3. Altere as opções aplicáveis. Opção Descrição Configuração de horário/data Ajusta o horário (formato 24 horas) e a data. comunicação Seleciona a taxa de transmissão (19200, 38400, 57600 ou 115200) e o protocolo (Modbus RTU ou ASCII) para a porta serial. Configurar base Seleciona a base do amostrador (Refrigerado para todos os climas, refrigerado, padrão portátil ou compacto portátil). Idioma Seleciona o idioma do controlador. Liq sensor cal (cal sensor de liq) Calibra o sensor líquido ou define como padrão. Ajuste de contraste Ajusta o contraste da tela de LCD. Ciclo de operação da bomba Altera o ciclo de operação da bomba. Faixa: de 50% a 100% (padrão 100%). 94 Português Opção Descrição Ajuste de senha Define uma senha para controlar o acesso aos menus de configuração do programa e do sistema. Selecione ENABLE>YES (Habilitar>Sim) para configurar uma nova senha ou ENABLE>NO (Habilitar>Não) para ativar a senha atual. Se a senha estiver sendo definida pela primeira vez, use 900900 como senha atual. Vida útil da tubulação Define um lembrete de quando a tubulação da bomba deve ser trocada. Vá até ENABLE>RESET CYCLES (Habilitar>Redefinir ciclos) para redefinir a contagem do ciclo da bomba como 0. Acesse ENABLE>CYCLE LIMIT (Habilitar>Limite do ciclo) para alterar o número de ciclos da bomba a ser usado para o lembrete. Vá até ENABLE>STATUS (Habilitar>Status) para mostrar o número atual de contagens de ciclo da bomba. SDI-12 (opcional) Configura a operação do dispositivo SDI–12. Essa opção é exibida somente quando um dispositivo é encontrado. Configure um programa de amostragem Um programa de amostragem inclui todos os parâmetros necessários para coletar amostras automaticamente. Colete as amostras em intervalos regulares ou quando o programa de amostragem estiver completo. 1. Pressione MENU. 2. Acesse PROGRAM SETUP>MODIFY ALL(Configuração do programa>Modificar tudo). A primeira tela de configuração será exibida. 3. Selecione ou insira valores para cada parâmetro. Use as teclas de seta para a esquerda e para a direita para mover para a esquerda ou direita. Use as teclas de seta para cima e para baixo para mudar um valor. Mude as unidades se necessário (gal. Ou mL) com as teclas de flechas. Opção Descrição garrafas Quantidade de garrafa: o número de garrafas do amostrador (1, 2, 4, 8 ou 24). Válvula de GRF cheia: ativada/desativada. Quando ativada, o controlador verifica a garrafa quanto às condições de transbordamento. Volume da garrafa: a capacidade de volume de cada garrafa. Faixa: 50–65000 mL (0,01 - 17,17 gal) tubulação de entrada Comprimento da tubulação: o comprimento da tubulação de entrada do filtro até o sensor de líquido. Faixa: 100–3000 cm ou 3–99 pés. Um comprimento preciso é necessário para obter um volume de amostra preciso. Tipo de tubo de entrada: o diâmetro e o material da tubulação de entrada (¼ pol. vinil, 3/8 pol. vinil ou 3/8 pol. Teflon). Atraso do programa Ritmo de amostra/coleta Ativado/desativado: quando ativado, o programa de amostragem inicia em um horário ou número de contagem especificado. Data e hora: a data e a hora em que o programa será iniciado (formato 24 horas). Contagens: o número de contagens ou pulsos de um medidor de fluxo no qual o programa será iniciado (1 a 9999 contagens). Se o ritmo de amostra for depois selecionado para horário, o atraso do programa será desabilitado. Baseado tempo—Consulte Coleção baseada em tempo na página 96. Observação: Quando CTVV é usado: • • • • • O sensor de líquido está sempre habilitado. A distribuição de amostras é composta. O modo de execução não é contínuo. A primeira amostra é obtida após o intervalo. A fonte de fluxo é sempre uma entrada de 4 a 20 mA. Baseado fluxo—Consulte Coleção baseada em fluxo na página 97. Português 95 Opção Descrição Distribuição de amostra Entregar amostras a todos>YES (Sim) Cada amostra é entregue a todas as garrafas. Finalizar após última: o programa para após a última amostra (1–999). Contínuo: o programa continua até ser parado manualmente. Entregar amostras a todos>NO (Não) As amostras são entregues a um subconjunto de garrafas . Amostras/garrafa: o número de amostras a ser coletado em cada garrafa (1–999). Garrafas/amostra: o número de garrafas que conterão a mesma amostra. sensor de líquido Habilitar ou desabilitar. Se desabilitado, o volume de amostra deverá ser calibrado pelo tempo. Volume de amostra O volume de cada amostra em ml (1 gal. = 3785,4 ml). Se for usado o modo de garrafas/amostras, cada garrafa recebe um volume completo de amostras (10–10.000 ml). Certifique-se de que o volume de amostra não exceda o volume de garrafas. Os volumes de amostra são arredondados para os 10 mL mais próximos. Lavagens de entrada O número de lavagens do tubo de entrada antes de uma amostra ser coletada (0–3). Tentativas de amostra O número de tentativas de amostragem após a falha de cada amostragem (0–3). Identidade do local O nome do local de amostragem (até 12 caracteres). A identidade do lugar será usada como um nome pré-definido se o programa for armazenado. amostragem avançada Vai até o menu de amostragem avançada. Se necessário, configure as opções no menu Advanced Sampling (Amostragem Avançada) menu. Executar programa A criação do programa de amostragem básica é concluída. O amostrador solicita ao usuário para iniciar ou cancelar o programa. Coleção baseada em tempo 1. Selecione ou insira os valores para cada opção. Opção Descrição Intervalo de ritmo Coleta amostras em intervalos de tempo regulares, em horas e minutos (0:01–999:00). Tome a primeira amostra O programa é iniciado imediatamente ou após o primeiro intervalo de tempo. Volume variável Especifique se os volumes de amostra variarão (baseados em fluxo). Em caso afirmativo, consulte Use volume variável na página 96. Em caso negativo, consulte Não use volume variável na página 97. Use volume variável A recolha de amostra é volume variável de tempo constante (CTVV). 1. Selecione ou insira os valores para cada opção. Opção Descrição Selecionar unidade Selecione as unidades de fluxo a serem usadas. Entrada Map 4 mA Insira a taxa de fluxo em concordância com a entrada de 4 mA. Entrada Map 20 mA Insira a taxa de fluxo em concordância com a entrada de 20 mA. Taxa de fluxo média Especifique a taxa de fluxo média para o site (1–999.999). As unidades são especificadas pelas unidades selecionadas acima. 96 Português Opção Descrição Volume de tempo desejado Especifique o volume total a ser coletado durante o programa de amostra (10–10.000 ml). Período de coleta Especifique o período de coleta total em horas e minutos (0:01–999:00). Volume mínimo da amostra Se o volume de fluxo não for suficiente para coletar este volume mínimo, a amostra será ignorada (10–10.000 ml). Não use volume variável 1. Selecione ou insira os valores para a opção. Opção Descrição Tome a primeira amostra Selecione se o programa iniciará imediatamente ou após ter passado o primeiro intervalo de tempo. Coleção baseada em fluxo Especifique a fonte de fluxo para amostragem baseada em fluxo: • Contagens—Consulte Use contagens para a fonte de fluxo na página 97. • Entrada de 4 a 20 mA—Consulte Use entrada de 4 a 20 mA para a fonte de fluxo na página 97. Use contagens para a fonte de fluxo 1. Selecione ou insira os valores para cada opção. Opção Descrição Tomar amostra cada Colete amostras a intervalos de fluxo regulares, em contagens (1–9.999 contagens). Tempo limite Colete uma amostra se o volume de fluxo for muito baixo. Hora O tempo máximo entre as amostras (0:01–999:00). O temporizador começa novamente após a coleta de cada amostra. Tome a primeira amostra O programa é iniciado imediatamente ou após o primeiro intervalo de tempo. Use entrada de 4 a 20 mA para a fonte de fluxo 1. Selecione ou insira os valores para cada opção. Opção Descrição Selecionar unidade Seleciona as unidades de fluxo a serem usadas. Entrada Map 4 mA Insira a taxa de fluxo em concordância com a entrada de 4 mA. Entrada Map 20 mA Insira a taxa de fluxo em concordância com a entrada de 20 mA. Tomar amostra cada Insira o fluxo total após o qual a amostra deve ser extraída. Cancelar tempo Selecione habilitar para forçar que uma amostra seja coletada se o volume de fluxo estiver excepcionalmente baixo. Português 97 Opção Descrição Hora O tempo máximo entre as amostras (0:01–999:00). O temporizador começa novamente após a coleta de cada amostra. Volume de amostra Insira o volume a ser coletado por amostra. Se o modo de garrafas/amostras for selecionado, cada garrafa recebe um volume completo de amostras (100–10.000 mL). Esta opção é ignorada quando o volume variável é selecionado. Como restaurar as configurações padrão A opção de restauração define todos os parâmetros de programa com as configurações padrão. 1. Pressione MENU. 2. Vá até PROGRAM SETUP>RESTORE (Configuração do programa>Restaurar). 3. Selecione YES (Sim). Iniciar ou parar um programa Use a tecla RUN/HALT PROGRAM (Executar/interromper programa) para iniciar ou parar um programa. Podem ser definidos pelo usuário até 12 vezes/datas de iniciar/parar. Um programa deve ser parado antes de ser possível modificar o programa ou a configuração do sistema. 1. Pressione RUN/HALT PROGRAM (Executar/interromper programa) 2. Selecionar uma das opções de saída. Opção Descrição Botão de inicialização Inicia o programa carregado atualmente. O estado muda para RUNNING (Em execução). Interromper Para o programa temporariamente. O estado muda para HALTED (Interrompido). Reiniciar O programa reinicia a partir do ponto no qual foi interrompido. O estado muda para RUNNING (Em execução). Começar novamente O programa inicia do princípio. O estado muda para RUNNING (Em execução). Finalizar programa Interrompe o programa atualmente em execução. O estado do programa muda para COMPLETE (Concluído). operação manual Use a operação manual para coletar uma amostra de tanque, mover o braço distribuidor, ou operar a bomba. Para obter mais informações, consulte a versão completa deste manual. Calibração de volume O volume de amostra pode ser calibrado automaticamente com o sensor de líquido ou manualmente. Quando um sensor de líquido for usado para calibragem de volume, o sensor será ajustado para medir com precisão todos os volumes programados. Se o sensor de líquido estiver desabilitado, o volume para amostras, a lavagens e água pluvial devem ser calibrados manualmente. Para aplicações típicas, pode ser usada água de torneira para a calibração de volume. Se a composição de amostra for significantemente diferente das amostras de água típicas, calibre o amostrador com o líquido a ser coletado. Todos os programas devem ser parados antes da calibração. Calibre o sensor de líquido A precisão do sensor de líquido varia com o tipo de líquido que está sendo amostrado. Por exemplo, o volume de uma amostra altamente túrbida poderá não ser tão preciso como aquele de água limpa. O sensor de líquido pode ser calibrado com água de amostra para aumentar a precisão de volume. 98 Português 1. Coloque a tubulação de entrada na água de amostra. 2. Pressione MENU. 3. Vá até PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuração do programa>Modificar selecionado(s)>Sensor de líquido>Habilitar) para garantir que o sensor de líquido esteja habilitado. 4. Acesse SYSTEM SETUP>LIQ SENSOR CAL (Configuração do sistema>Cal. do sensor de líquido). 5. Selecione PERFORM CAL (Executar calibração). 6. Selecione START (Iniciar). A bomba opera em reverso para purgar o tubo de entrada. Em seguida, a bomba opera na direção para frente. 7. Observe o fluxo da amostra no tubo de entrada para verificar se a amostra está passando pelo sensor de líquido. 8. Quando a amostra estiver passando pelo sensor de líquido e não houver a presença de bolhas, pressione STOP (Parar). O tubo de entrada será automaticamente purgado. A calibração sensor de líquido está concluída. 9. Selecione DONE (Concluído) para sair e REPEAT (Repetir) para fazer a calibração novamente. Como restaurar a calibração padrão É possível voltar a calibração de volume do sensor de líquido para as configurações padrão de fábrica. 1. Selecione PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuração do programa>Modificar selecionado(s)>Sensor de líquido>Habilitar) para garantir que o sensor de líquido esteja habilitado. 2. Pressione VOLUME CALIBRATION (Calibração de volume) e selecione RESET CAL (Redefinir calibração). A calibração de volume será definida com as configurações padrão. Será mostrada a mensagem “value has been reset to 0” (valor redefinido como 0). Calibre o volume de amostra com o sensor de líquido Faça a calibração do volume de amostra com o sensor de líquido habilitado para ajustar levemente os volumes de amostra. 1. Vá até PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>ENABLE (Configuração do programa>Modificar selecionado(s)>Sensor de líquido>Habilitar) para garantir que o sensor de líquido esteja habilitado. 2. Coloque a tubulação de entrada na água de amostra ou em água de torneira. Observação: Use a água de amostra para obter uma precisão melhor. 3. Coloque tubulação da saída da bomba em um cilindro graduado. 4. Pressione VOLUME CALIBRATION (Calibração de volume) e selecione CALIBRATION (Calibração). 5. Selecione START (Iniciar). A bomba opera em reverso para purgar o tubo de entrada. Em seguida, a bomba opera na direção para frente e coleta o volume de amostra especificado no programa de amostragem. A bomba opera em reverso para purgar o tubo de entrada. 6. Quando a amostra tiver sido coletada, selecione DONE (Concluído). Compare o volume da amostra no cilindro com graduação com o volume de amostra que está inserido no menu de configuração do programa. 7. Se o volume que foi coletado for diferente do volume de amostra no programa atual, insira o volume que foi realmente coletado. Pressione ENTER. O sensor é ajustado para medir com precisão todos os volumes programados. 8. Selecione START (Iniciar) para iniciar o programa de amostragem ou CANCEL (Cancelar) para sair do menu de calibração do volume. Português 99 Calibre o volume de amostra manualmente Quando o sensor de líquido estiver desabilitado, todos os volumes deverão ser calibrados manualmente por tempo. O volume de amostra está calibrado para o volume especificado no programa atual. Se o volume de amostra for mudado no programa, o volume de amostra deverá ser calibrado novamente para o novo volume. 1. Selecione PROGRAM SETUP>MODIFY SELECTED>LIQUID SENSOR>DISABLE (Configuração do programa>Modificar selecionado(s)>Sensor de líquido>Desabilitar) para garantir que o sensor de líquido esteja desabilitado. 2. Coloque a tubulação de entrada na água de amostra ou em água de torneira. 3. Coloque tubulação da saída da bomba em um cilindro graduado. 4. Pressione VOLUME CALIBRATION (Calibração de volume). Se for exibido mais de um volume, selecione um dos volumes para a calibração. 5. Selecione START (Iniciar). A bomba opera em reverso para purgar o tubo de entrada. A bomba então opera na direção para frente e começa a coletar amostras. 6. Pare a bomba no volume selecionado ou local de lavagem: • Volume de amostra: pressione STOP (Parar) quando o volume especificado no programa principal for coletado. • Enxágue: pressione STOP (Parar) quando o líquido na tubulação de entrada alcançar o detector de líquido. • Volume pluvial: pressione STOP (Parar) quando o volume especificado no programa águas pluviais for coletado. 7. Selecione DONE (Concluído) para ir para a calibração do próximo volume ou REPEAT (Repetir) para fazer a calibração novamente. 8. Quando todos os volumes tiverem sido calibrados, selecione DONE (Concluído). 9. Conecte a tubulação de saída ao encaixe de tubo no amostrador. 10. Selecione START (Iniciar) para iniciar o programa de amostragem ou CANCEL (Cancelar) para sair. Verifique o volume de amostra Para garantir que o volume de amostra esteja preciso, colha uma amostra rápida. Não volte na calibragem para verificar o volume já que a compensação de volume está reajustada em zero no começo de uma calibragem. 1. 2. 3. 4. 5. 6. 7. 8. Pressione MANUAL OPERATION (Operação manual). Selecione GRAB SAMPLE (Amostra rápida). Coloque a tubulação de entrada na água de amostra. Coloque tubulação da saída da bomba em um cilindro graduado. Se o sensor de líquido estiver habilitado, insira o volume a ser verificado. Se o sensor de líquido estiver desabilitado, insira o volume que está especificado no programa. Pressione ENTER. O ciclo da bomba é iniciado (purgação-amostra-purgação). Compare o volume que foi coletado no cilindro graduado com o volume esperado. Se o volume coletado não estiver correto, faça novamente a calibração de volume. Manutenção PERIGO Vários perigos. Somente pessoal qualificado deve realizar as tarefas descritas nesta seção do manual. 100 Português PERIGO Risco de eletrocussão. Retire o instrumento da energia antes de realizar atividades de manutenção ou serviço. ADVERTÊNCIA Exposição a riscos biológicos. Siga os protocolos de manuseio com segurança durante o contato com os frascos de amostras e componentes do amostrador. ADVERTÊNCIA Vários perigos. O técnico deve garantir que o equipamento funcione segura e corretamente após os procedimentos de manutenção. Como limpar o instrumento CUIDADO Perigo de incêndio. Não use agentes inflamáveis para limpar o instrumento. AVISO Não limpe o aquecedor do compartimento do controlador com líquidos de qualquer natureza. Se a água não for suficiente para limpar o controlador e a bomba, desconecte o controlador e movao para longe do amostrador. Aguarde tempo suficiente para que o controlador e a bomba sequem antes de reinstalar as peças e retomar os serviços. Limpe o amostrador da seguinte maneira: • Gabinete do amostrador — limpe as superfícies internas e externas com um pano úmido e detergente suave. Não use limpadores ou solventes abrasivos. • Garrafas de amostra — limpe as garrafas com uma água e escova e detergente suave. Enxágue com água limpa e, em seguida, água destilada. Use uma autoclave para limpar as garrafas de vidro, se necessário. Substituir o dessecante Um cartucho dessecante no controlador absorve a umidade e evita a corrosão. Monitore a cor do dessecante através do visor (Figura 12). O dessecante em boas condições é amarelo. Quando sua estiver verde, substitua o dessecante. 1. 2. 3. 4. 5. 6. Desatarraxe e remova o cartucho dessecante (Figura 12). Remova o plugue e descarte o dessecante gasto. Encha o tubo dessecante com dessecante fresco. Instale o plugue. Aplique graxa ao O-ring. Instale o tubo dessecante no controlador. Português 101 Figura 12 Cartucho dessecante 1 Plugue 3 O-ring 2 Tubo de dessecante 4 Visor do dessecante Manutenção da bomba CUIDADO Risco de pressionamento. Desconecte o instrumento da alimentação antes de realizar atividades de manutenção ou serviço. Como substituir a tubulação da bomba AVISO O uso de tubulação diferente daquela fornecida pelo fabricante poderá causar desgaste excessivo das peças mecânicas e/ou desempenho ruim da bomba. Examine a tubulação da bomba para ver se há desgaste no local em que os cilindro entram em fricção com a tubulação. Substitua a tubulação quando a mensagem de lembrete for exibida ou quando a tubulação mostrar sinais de desgaste. Pré-requisitos: • Tubulação da bomba: pré-cortada ou longa de 4,6 m ou 15,2 m (15 pés ou 50 pés) 1. Desconecte a energia do controlador. 2. Se for usada a tubulação longa, corte a tubulação e adicione pontos de alinhamento. Consulte o manual expandido no site do fabricante. 3. Remova a tubulação da bomba como mostrado nas etapas ilustradas que seguem. 4. Limpe o resíduo de silicone do interior do alojamento da bomba e dos cilindros. 5. Instale a nova a tubulação da bomba como mostrado nas etapas ilustradas que seguem. 102 Português Português 103 Como limpar o rotor Limpe o rotor, os trilhos do tubo da bomba e o alojamento da bomba com detergente suave. Consulte Como substituir a tubulação da bomba na página 102 e as etapas ilustradas seguintes. Como substituir o tubo do braço distribuidor O braço distribuidor se move sobre cada garrafa durante amostragem de garrafas múltiplas. Substitua o tubo do braço do distribuidor quanto o tubo estiver gasto. Certifique-se de que o tubo correto é usado para o distribuidor correto e o braço distribuidor. Observação: A tubulação do distribuidor não é a mesma que a tubulação da bomba. A tubulação da bomba instalada no conjunto do distribuidor pode danificar o distribuidor. Além disso, as amostras podem ser perdidas, pois o braço de distribuidor não se move com facilidade. 1. Remova o tubo do braço do distribuidor e do teto da do gabinete do amostrador. 2. Insira o novo tubo no braço do distribuidor. Certifique-se de que o tubo esteja alinhado com as dimensões indicadas em Figura 13. 3. Insira a outra extremidade do tubo no encaixe do teto da do gabinete do amostrador. 4. Faça o teste de diagnóstico do distribuidor para garantir que está funcionando corretamente. Figura 13 Conjunto do distribuidor 1 Bico 3 Motor distribuidor 2 Braço distribuidor 4 Eixo 104 Português 5 Tubo do distribuidor Substituição de fusível O fusível não é um item que pode ser reparado pelo usuário. Entre em contato com o suporte técnico. Um fusível queimado pode ser uma indicação que o amostrador tem um problema e que o serviço é necessário. Português 105 目录 规格 第 106 操作 第 119 基本信息 第 107 维护 第 124 安装 第 110 附加信息 制造商网站中提供了附加信息 规格 产品规格如有变化,恕不另行通知。 SD900 便携式采样器 规格 标准底座 小型底座 复合底座 尺寸 直径:50.5 厘米(19.9 英寸) 直径:44.1 厘米(17.4 英寸) 高度:69.4 厘米(27.3 英寸) 高度:61 厘米(24 英寸) 直径:50.28 厘米 (19.8 英寸) 高度:79.75 厘米 (31.4 英寸) 重量 外壳 15 kg (35.6 lb),装 1-L 聚乙烯瓶 (24 个) 12.2 kg (27 lb),装 575-mL (19.44 oz) 聚乙烯瓶(24 个) 14.8 kg (32.6 lb),装 10-L (2.5 gal) 聚乙 烯容器(1 个) 12.9 kg (28.3 lb),装 10-L (2.5 gal) 聚乙烯容器(1 个) 15 kg (36 lb),装 950-mL (32.12 oz) 玻璃瓶 (12 个) 耐冲击 ABS,3 节构造;双壁底座,带 2.54 cm (1 in.)保温层 - 直身瓶,与冰直接接触。 样品温 0–60 °C (32–140 °F) 度 过滤器 标准尺寸、适合于较浅应用的高流速或低收缩 316 不锈钢、Teflon® 或标准尺寸 316 不锈钢 样品抽 9.5 mm (3/8 in.)I.D. 乙烯基或特氟龙衬里聚乙烯 取试管 采样瓶 1-L (0.26 gal) 聚乙烯和/或 350-mL 容量 (11.83 oz) 玻璃瓶(24 个) 575-mL (19.44 oz) 聚乙烯瓶 (24 个) 2.3-L (0.6 gal) 聚乙烯和/或 1.9-L (0.5 gal) 玻璃瓶(8 个) 950-mL (32.12 oz) 玻璃瓶 (8 个) 3.8-L (1 gal) 聚乙烯和/或 3.8-L (1 gal) 玻 璃瓶(4 个) 10-L (2.5 gal) 聚乙烯瓶(1 个) 10-L (2.5 gal) 玻璃瓶(1 个) 3.8-L (1 gal) 聚乙烯和/或 3.8-L (1 gal) 玻 璃瓶(2 个) 21-L (5.5 gal) 或 15-L (4 gal) 聚乙烯复合 容器或 20-L (5.25 gal) 聚乙烯或 10-L (2.5 gal) 聚乙烯或 10-L (2.5 gal) 玻璃瓶 (1 个) SD 控制器 规格 详细信息 尺寸 (W x H x D) 29.2 x 17.1 x 26.4 cm (11½ x 6¾ x 103/8 in.) 重量 4.2 kg (9.26 lb) 外壳 PC/ABS 复合材料,NEMA 4X,6,IP 67 106 中文 21-L (5.5 gal) 聚乙烯瓶 (1 个) 规格 详细信息 电源要求 由一个 8754500 电源供应 15 V 直流电; 过载保护 泵用 7 A 直流线路保险丝 操作温度 0 到 50 °C(32 到 122 °F) 存储温度 –30 至 60 °C (–22 至 140 °F) 存放/操作湿度 100% 冷凝 泵 高速蠕动泵,带有弹簧安装式尼拉特隆滚柱 泵外壳 IP37 泵管 9.5 mm(内径)x 15.9 mm(外径)(3/8 in.(内径)x 5/8) in.(外径))硅树脂管 泵管寿命 20000 次采样循环:1 L (0.3 gal) 采样量,1 次冲洗,6 分钟步调间隔,4.9 m (16 ft) 3/8 in.进样管,4.6 m (15 ft) 垂直升程,21 °C (70 °F) 采样温度 垂直采样升程 最大 8.5 m (28 ft):8.8 m (29 ft) 3/8 in.乙烯基进样管处于海平面高度且温度为 20–25 °C (68–77 °F) 泵流速 4.8 L/min (1.25 gpm),其中垂直升程为 1 m (3 ft),并且采用 3/8 in.典型进样管 采样量 能以 10 mL (0.34 oz) 程控间隔从 10 mL 递增至 10000 mL (从 3.38 oz 递增到 2.6 gal) 样品量重复性(典型 值) ±5% × 200 mL 采样量,其中使用未经校准的液样检测方式:4.6 m (15 ft) 垂直升程, 4.9 m (16 ft) 3/8 in.乙烯基进样管,单瓶,满瓶关闭,室温,海拔 1524 m (5000 ft) 采样量精度(典型值) ±10% × 200 mL 进样量,其中使用未经校准的液样检测方式:4.6 m (15 ft)垂直升程, 4.9 m (16 ft) 3/8 in.乙烯基进样管,单瓶,满瓶关闭,室温,海拔 1524 m (5000 ft) 采样模式 步调:固定时间,固定流量,可变时间,可变流量,常数时间可变样品量 (CTVV)。 请参阅本手册扩展版本中的常数时间可变样品量 (CTVV) 采样。 分布:单瓶复合、多瓶复合、多瓶分离、每个样品有多瓶、每瓶有多个样品、每个样品 有多瓶、每瓶有多个样品 运行模式 按照用户输入的采样次数连续运行或断续运行 多个程序 存储至多三个采样程序 输送速率(典型值) 0.9 m/s (2.9 ft/s):4.6 m (15 ft) 垂直升程,4.9 m (16 ft) 3/8 in.乙烯基进样管,温度 21 °C (70 °F),海拔 1524 m (5000 ft) 液样传感器 超声波式。主体:通过了 Ultem NSF ANSI 标准 51,符合 USP Class VI 采样历史记录 最多 510 个记录 ® 空气吹扫 每次采样前后自动进行空气吹扫。持续时间将会自动补偿进样管长度的变化。 事件日志 510 条记录 连接 电源连接、辅助连接、串行通信连接、分配器连接、SDI-12 过流材料 不锈钢、聚乙烯、聚四氟乙烯、聚醚酰亚胺、硅树脂 通信 RS232, Modbus, SDI-12 保修期 1年 基本信息 对于因本手册中的任何不足或遗漏造成的直接、间接、特别、附带或结果性损失,制造商概不负责。 制造商保留随时更改本手册和手册中描述的产品的权利,如有更改恕不另行通知或承担有关责任。修 订版可在制造商的网站上找到。 中文 107 安全信息 注意 对于误用和滥用造成的产品损坏,制造商概不负责,包括但不限于:直接、附带和间接的损坏,并且对于适用法律 允许的最大程度的损坏也不承担任何责任。用户唯一的责任是识别重大应用风险和安装适当的系统,以在设备可能 出现故障时保护工艺。 请在拆开本设备包装、安装或使用本设备前,完整阅读本手册。特别要注意所有的危险警告和注意事 项。否则,可能会对操作者造成严重的人身伤害,或者对设备造成损坏。 确保设备提供的保护没有受损。请勿以本手册指定方式之外的其它方式使用或安装本设备。 危险信息使用 危险 表示潜在的或紧急的危险情况,如果不加以避免,将会导致死亡或严重伤害。 警告 表示潜在或非常危险的情形,如不避免,可能导致严重的人身伤亡。 警告 表示潜在的危险情形,可能导致一定程度的人身伤害。 注意 表明如不加以避免则会导致仪器损坏的情况。需要特别强调的信息。 警告标签 请阅读贴在仪器上的所有标签和标记。 如未遵照这些安全标签的指示操作,则可能造成人身伤害或仪 器损坏。 仪器上的符号在手册中通过警告说明参考。 这是安全警报标志。请遵守此标志后面的所有安全信息,以避免可能造成的伤害。如果仪器上有此标 志,则请参见仪器手册,了解操作或安全信息。 此标记指示存在潜在的夹伤危险。 标有此符号的电气设备在欧洲不能通过家庭或公共垃圾系统进行处理。请将老旧或报废设备寄回至制造 商处进行处置,用户无需承担费用。 认证 加拿大无线电干扰产生设备法规(Canadian Radio Interference-Causing Equipment Regulation),IECS-003,A 类: 制造商支持测试记录留存。 此 A 类数字设备符合加拿大干扰产生设备法规的所有要求。 Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences. FCC 第 15 部分,“A”类限制 制造商支持测试记录留存。该设备符合 FCC 规定第 15 部分的要求。设备操作满足以下两个条件: 1. 本设备不会造成有害干扰。 2. 本设备必须接受任何接收到的干扰,包括可能导致意外操作的干扰。 108 中文 若未经负责出具符合声明的一方明确同意擅自对本设备进行改动或改装,可能会导致取消用户操作该 设备的权限。本设备已经过测试,符合 FCC 规定第 15 部分中确定的 A 类数字设备限制。这些限制专 门提供当设备在商业环境下工作时针对有害干扰的合理保护。该设备产生、使用和放射无线电射频能 量,如果不按照说明手册的要求对其进行安装和使用,可能会对无线电通讯造成有害干扰。本设备在 居民区工作时可能会产生有害干扰,这种情况下用户须自行承担费用消除这种干扰。以下方法可用于 减少干扰问题: 1. 2. 3. 4. 5. 断开设备的电源,以便确证它是干扰源与否。 如果设备与遭受干扰的仪器连接到相同的插座,将设备连接到其他插座。 将设备从接受干扰的仪器边上移开。 重新定位受干扰仪器的接收天线。 同时尝试以上多项措施。 产品概述 危险 化学或生物危害。如果该仪器用于监测处理过程和/或监测有法规限制以及有与公众健康、公众安全、食 品或饮料制造或加工相关监测要求的化学药物添加系统时,仪器的使用者有责任了解并遵守所有适用的 法规,并且要建立适当的机制,确保在仪器发生故障的时候也不会违法这些法规。 警告 火灾危险。此产品不得与易燃液体一同使用。 SD900 便携式采样器按指定的间隔时间采集液体样本,并将样品存放在采样瓶或容器中。该采样器适 用于各种水体应用以及有毒污染物和悬浮物。安装采样器和各种固定座、采样瓶或容器。请参阅 图 1。 图 1 产品概述 1 小型底座 6 控制器接线 2 标准绝缘底座 7 泵 3 中心部分 8 控制器 4 电源 9 液样传感器 5 顶盖 10 标准绝缘底座,适合 21-L (5.5 gal) 容器 产品组件 确保已收到所有组件。 请参阅 图 2。 如有任何物品丢失或损坏,请立即联系制造商或销售代表。 中文 109 图 2 产品组件 1 底座(标准型、小型或复合型) 6 电池充电器(选件) 2 单瓶安装的组件(瓶子和支架可更换) 7 交流电源(可选) 3 中心部分(带控制器) 8 过滤器 4 电池(选件) 9 进样管,乙烯基或特氟龙衬里 5 顶盖 10 多瓶安装的组件(瓶子和固定座可更换) 安装 危险 多种危险。只有合规的专业人员才能从事文档中本部分所述的任务。 密闭空间预防措施 危险 爆炸危险。在进入密闭空间之前,需要进行有关进入之前测试、通风、进入过程、疏散救援过程以及安 全工作做法的培训。 下列信息旨在帮助用户了解进入密闭空间时可能遇到的危险和风险。 1993 年 4 月 15 日,OSHA 最终颁布法律规定 CFR 1910.146,密闭空间作业许可 (Permit Required Confined Spaces)。该标准对美国 250000 多个工业场地具有直接影响,用于保护狭窄空间内工人的 健康和安全。 密闭空间定义: 密闭空间是指具有下列一种或多种情况(或具有直接可能性)的任何位置或壳体: • 氧气浓度低于 19.5% 或高于 23.5% 并且/或者硫化氢 (H2S) 浓度高于 10 ppm 的空气。 • 含有易燃易爆的气体、蒸气、雾气、粉尘或纤维的空气。 • 有毒材料,接触或吸入时会引起受伤、损害健康或致死。 密闭空间不适用于人类居住。密闭空间限制进入,其中含有已知的或潜在的危害。密闭空间包括检修 孔、堆栈、管道、大桶、开关地下室及其他类似位置。 进入可能含有危险气体、蒸气、雾气、粉尘或纤维的密闭空间和/或位置之前,务必遵守标准的安全规 程。进入密闭空间之前,请首先查找并阅读所有相关的规程。 110 中文 机械安装 现场安装指导 危险 爆炸危险。该仪器未获批准在危险场所安装。 请参阅现场位置评估指南。 • 如果采样器的安装空间有限,请遵循所有安全保护措施。请参阅 密闭空间预防措施 第 110。 • 确保该位置的温度在规范范围内。请参阅 规格 第 106。 • 将采样器安装在平面上或使用悬架装置、支架或拉紧杆悬挂采样器。请参阅 将采样器安装在检修孔 中 第 111 和适用的安装文档。 • 尽可能接近样品源以降低分析延迟。请参阅 使采样器保持垂直 第 114。 • 有关载运速度和最大垂直扬程的限制信息,请参阅 规格 第 106。 将采样器安装在检修孔中 将采样器安装在检修孔中样水上方。为采样器安装拉紧杆或支架。将拉紧杆安装在检修孔内部。拉紧 杆受压支固在侧壁上。支架的宽度与检修孔护盖宽度相同。在支架护盖的正下方安装支架。请参阅制 造商网站上提供扩展版手册。请参阅随附件提供的文档安装采样器。 安装分配器或满瓶关断装置(选件) 分配器或满瓶关断装置组件一般出厂时已安装。请参阅分配器或满瓶关断装置文档了解安装说明。 准备采样器 清洁采样瓶 用刷子、水及温和的清洁剂清洁采样瓶和瓶盖。先后用淡水及蒸馏水冲洗容器。 安装一个采样瓶 利用一个采样瓶采集一种混合样品。采样瓶满样时,满瓶关闭功能将会停止采样程序。 1. 清洁采样瓶。请参阅 清洁采样瓶 第 111。 2. 按照 图 3 所示安装采样瓶。 中文 111 图 3 单瓶安装 1 聚乙烯瓶,10 L (2.5 gal) (1918) 6 聚乙烯瓶,21 L (5.5 gal) (6494) 2 玻璃瓶,10 L (2.5 gal) (6559) 7 小型底座 (8975) 3 支架 (1502) 8 标准绝缘底座 (8976) 4 聚乙烯瓶,15 L (4 gal) (1367) 9 标准绝缘底座,适合 21 L (5.5 gal) 的采样瓶 (8561) 5 聚乙烯瓶,19 L (5 gal) (6498) 为多个采样安装第一个采样瓶 用多个采集瓶采集样品,装入单独的采集瓶或多个采集瓶。分配器将采样管移至每个采样瓶上方。按 照 图 4 中所示将采样瓶安装在采样器底座中。将第一个采样瓶(1 号)安装在采样器底座标签下方。 根据标签指示的方向,按升序逐号安装其余的采样瓶。请参阅 图 5 第 114 查看必要的组件图。 112 中文 图 4 1 号瓶安装 1 1 号瓶安装位置(小型底座) 6 弹力带 2 1 号瓶位置,24 个瓶 7 标准绝缘底座,适合 21 L (5.5 gal) 的采样瓶 3 1 号瓶位置,8 个瓶 8 标准绝缘底座 4 1 号瓶位置,2 个或 4 个瓶 9 小型底座 5 固定座 安装多个采样瓶 安装多个采样瓶时,分配臂会使采样管移动到每个瓶子的上方。采集的样品达到指定数量时,自动结 束采样。 1. 清洁采样瓶。请参阅 清洁采样瓶 第 111。 2. 在采样器底座标签下方安装第一个采样瓶(1 号)请参阅 为多个采样安装第一个采样瓶 第 112。 3. 按照 图 5 所示装配采样瓶。对于 8 个或更多采样瓶,请确保第一个采样瓶以顺时针方向紧靠第一 个瓶符进行放置。 4. 将采样瓶组件放在采样器中。对于 8 个或更多采样瓶,对齐底部瓶托槽中的接线。 中文 113 图 5 多瓶安装 1 定位座/固定座 (2347) 8 玻璃瓶,3.8 L (1 gal)(4 个)(2216),或聚乙烯瓶, 3.8 L (1 gal)(4 个)(2217) 2 固定座 (1422) 9 玻璃瓶,1.9 L (0.5 gal)(8 个)(1118),或聚乙烯 瓶,2.3 L (0.61 gal),(8 个)(657) 3 固定座 (2190) 10 聚乙烯瓶,1 L (0.26 gal)(24 个)(737) 4 固定座 (2189) 11 玻璃瓶,350 mL (11.83 oz)(24 个)(732) 5 玻璃瓶,950 mL (32.12 oz)(8 个)(2348) 12 小型底座 (8975) 6 聚乙烯瓶,575 mL (19.44 oz)(24 个)(1369) 13 标准底座 (8976) 7 玻璃瓶,3.8 L (1 gal)(2 个)(2214),或聚乙烯瓶, 3.8 L (1 gal)(2 个)(2215) 使采样器保持垂直 在采样流中间部位安装过滤器(不要在表面或底部安装),确保采集到具有代表性的样品。请参阅 图 6。 请参阅 图 7 了解进样管的安装。 1. 按照 图 7 所示将进样管连接到采样器。 注: 如果使用特氟龙衬里的管子,则采用连接套件 2186。 2. 将进样管和过滤器安装在样品源的主液流中,即水处于湍流状态并且混合均匀的位置。 • 使进样管尽可能短。 • 使进样管保持最大的垂直斜度,以便在两次采样期间完全排空管子。 注: 如果无法保持垂直倾斜,或是管子加压,则禁用液体传感器。手动校准采样量. 请参阅 手动校准采样量 第 124。 • 确保进样管未扭曲。 114 中文 图 6 仪器安装 1 过滤器 3 进样管 2 垂直上升 图 7 进样管安装 电气安装 控制器接线 警告 电击危险。外部连接的设备必须具有适用的国家安全标准评估。 图 8 所示是控制器的接线。 中文 115 图 8 控制器接头 1 辅助装置 4 SDI-12 装置选项 2 电源 5 分配器/满瓶关闭 3 串行通信 连接一个流量计 将一个流量计连接到控制器,以便在样品流量高于或低于指定值时启动或关闭采样器。 需收集的物品: • Sigma 流量计用多功能全电缆(或 980 型流量计用 980 半电缆)。 • 可选分接头增加连接。两个或多个分接头可以串行连接。 1. 将电缆一端连接至流量计。对于 980 型流量计,请参阅 980 型流量计用户手册。 2. 将电缆的另一端连接到控制器的辅助装置接头上。 注: 如果流量计采用一根 6 引脚电缆,则使用 6 引脚转 7 引脚适配电缆。 连接一台个人电脑或通信网络 将一台个人电脑或通信网络连接到控制器,以便传送数据或配置采样器。 需收集的物品: • 串行电缆,7 引脚,由 RS232 接口转接到 DB-9 接口(请参阅 图 9)。 接线:B 接 5(信号接 地);D 接 3 (RCD);F 接 2 (TXD);G(接地)。 1. 将串行电缆的一端连接到通信设备或网络上。 2. 将电缆的另一端连接到控制器的串行通信接头上。 图 9 串行通信电缆 将采样器连接至电源 危险 电击致命危险。如果此设备在户外或在可能潮湿的场所使用,则必须使用接地故障电路中断器 (GFCI/GFI) 将此设备连接到其主电源。 116 中文 危险 火灾危险。在电源线中安装 15 A 断路器。断路器如果极为贴近设备,则可以是本地电源断开装置。 危险 电击致命危险。需要连接保护接地地线。 警告 电击致命危险。确保可轻松切断本地电源。 警告 电击危险。如果泵循环之间的时间太短,则电源可能温度过高。 注意 确保泵的循环时间不会造成电源过热。请参阅 计算泵的循环时间 第 117。 使用电池或交流电源为采样器供电。请参阅下面所示的说明步骤。 计算泵的循环时间 在两次采样循环中间,泵必须关闭至少一段时间,以便电源不会过热。 1. 计算泵在采样循环期间连续运行的总时间。包括所有阶段:清洗前、吸入清洗、采样、采样重试和 清洗后。 2. 根据图 10 确定两次采样循环期间泵保持关闭的最短时间。 3. 确保采样器程序的步调或时间间隔让泵在两次采样循环中间关闭至少一段时间。 中文 117 图 10 泵在 50 °C (122 ºF) 下启动和关闭的时间 1 泵必须关闭的时间(以分为单位) 2 泵连续运行的时间(以分为单位) 3 示例:如果泵连续运行 5 分钟,则泵必须关闭 15 分 钟。 开启电源 按下电源键接通或断开电源。确保电源(交流电源或电池供电)正确安装。 用户界面及导航 控制器键盘如 图 11 所示。利用 MENU(菜单键) 可以设置采样程序、配置控制器的设置或完成诊断 测试。利用箭头键、ENTER(输入键)及 BACK(返回键)可以滚动浏览菜单、进行选择及输入数 值。查看显示屏幕底部或边侧的箭头可以了解更多可用选项。 118 中文 图 11 控制器键盘 1 POWER(电源) 5 MENU(菜单) 9 STATUS(状态) 2 VOLUME CALIBRATION(采样 量校准) 6 ENTER(输入) 10 RUN/HALT PROGRAM(运 行/中止程序) 3 STOP(停止) 7 BACK(返回) 4 LED(LED 指示灯) 8 ARROW KEYS(箭头键) 11 MANUAL OPERATION(手动 操作) 操作 配置系统的设置 1. 按下 MENU(菜单)。 2. 进入 SYSTEM SETUP(系统设置)。 3. 更改可用选项。 选项 说明 Time/date setup(时间/日期设 置) 设置时间(24 小时格式) 和日期。 Communication(通信) 为串行端口选择波特率(19200、38400、57600 或 115200)和协议 (Modbus RTU 或 ASCII)。 Setup base(底座设置) 选择采样器底座(全天候冷藏式、冷藏式、便携式标准型或便携式紧凑 型)。 Language(语言) 选择控制器的语言。 Liq sensor cal(液样传感器校 准) 校准液样传感器或设为默认。 Set contrast(设置对比度) 调整 LCD 屏幕的对比度。 Pump duty cycle(泵的负载周 期) 更改泵的负载周期。周期范围:50%~100%(默认:100%)。 中文 119 选项 说明 Password setup(密码设置) 设置一个密码,以控制对程序设置菜单及系统设置菜单的访问。选择 ENABLE(启用)>YES(是)以设置一个新密码,或选择 ENABLE(启 用)>NO(否)以激活当前密码。如果首次设置密码,则使用 900900 作 为当前密码。 Tubing life(管子寿命) 设置一个泵管更换提示。进入 ENABLE(启用)>RESET CYCLES(重置 循环),将泵循环计数重置为 0。进入 ENABLE(启用)>CYCLE LIMIT (循环限制)以更改所提示的泵循环数。进入 ENABLE(启用) >STATUS(状态)以显示当前的泵循环计数。 SDI–12 (optional)(SDI–12(可 选)) 配置 SDI–12 装置的运行。只有在发现一个装置时,才会显示该选项。 设置一个采样程序 采样程序包括自动采样所需的所有参数。定期采样或在采样程序结束时进行采样。 1. 按下 MENU(菜单)。 2. 进入 PROGRAM SETUP(程序设置)>MODIFY ALL(更改所有项)。然后会显示第一个屏幕。 3. 为每个参数选择或输入数值。利用左右箭头键可左右移动。利用上下箭头键可更改数值。如有必 要,用箭头键修改单位(gal 或 mL)。 选项 说明 Bottles(采样 瓶) 采样瓶数量:采样器中的瓶子数量(1、2、4、8 或 24)。 满瓶关闭:启用/禁用。如果启用该功能,则控制器会检查采样瓶是否出现溢流状况。 瓶容量:每个采样瓶的体积容量。容量范围:50–65000 mL (0.01 - 17.17 gal) Intake tubing(进 管长:过滤器至液样传感器之间的进样管长度。长度范围:100–3000 cm 或 3–99 ft。 精 样管) 确的管长才能获得精确的采样量。 进样管类型:进样管的直径和材料(¼ in. 乙烯基塑料管、3/8 in. 乙烯基塑料管或 3/8 in. 特氟龙管)。 Program delay (程序延迟) Sample pacing/ collection(采样 步调/样品采集) 启用/禁用:选择“启用”时,采样程序会在指定时刻或计数下启动。日期和时间:程序启动 时的日期和时间(24 小时格式)。计数:程序将要启动时,流量计的计数或脉冲数(1~ 9999 个计数)。如果采样步调随后设为时间形式,则程序延迟功能将被禁用。 基于时间 - 请参阅 基于时间的采集 第 121。 注: 当使用 CTVV 时: • • • • • 始终启用液体传感器。 样品分布为复合式。 运行模式为非持续性。 间隔时间之后采集第一个样品。 流量源通常为 4–20 mA 输入。 基于流量 - 请参阅 基于流量的采集 第 121。 Sample distribution(样 品分布) Deliver samples to all(将样品输送到所有采样瓶)>YES(是) 每个样品均被输送到所有采样瓶。采样结束:最后一次采样之后结束程序 (1–999)。连续 采样:继续执行采样程序,直到手动停止为止。 Deliver samples to all(将样品输送到所有采样瓶)>NO(否) 将样品输送到一部分采样瓶中。样品/采样瓶:每个采样瓶需要采集的样品数量 (1–999)。 采样瓶/样品:含有相同样品的采样瓶的数量。 Liquid sensor (液样传感器) 启用或禁用。如禁用,采样量必须按时间校准。 Sample volume (采样量) 每个样品的采集量,单位为 mL (1 gal = 3785.4 mL)。如果使用采样瓶/样品模式,则每个 采样瓶将会采集满瓶样品 (10–10000 mL)。确保采样量不超过采集瓶体积。采样量以 10 mL 为基础四舍五入。 120 中文 选项 说明 Intake rinses(进 样冲洗) 采样之前冲洗的进样管数量 (0–3)。 Sample retries (采样重试次数) 采样失败之后尝试的采样次数 (0–3)。 Site ID(采样点 ID) 采样位置的名称(最长 12 个字符)。保存程序时用采样点 ID 作为预设文件名。 Advanced sampling(高级 采样) 进入高级采样菜单。必要时在 Advanced Sampling(高级采样)菜单中配置相应选项。 运行程序 基本的采样程序创建已完成。采样器将提示用户启动或取消采样程序。 基于时间的采集 1. 选择选项或为每个选项输入数值。 选项 说明 步调间隔 定时采样的时间间隔,以小时和分钟为单位 (0:01–999:00)。 采集第一个样品 程序立即启动,或在第一个间隔时间之后启动。 可变样品量 指定样品量是否可变(基于流量)。如果 yes(是),请参阅 使用可变样品量 第 121。如果 no(否),请参阅 切勿使用可变样品量 第 121。 使用可变样品量 样品步调为常数时间可变样品量 (CTVV)。 1. 选择选项或为每个选项输入数值。 选项 说明 选择单位 选择要使用的流量单位。 设定 4mA 输入 输入与 4mA 输入相一致的流量。 设定 20mA 输入 输入与 20mA 输入相一致的流量。 平均流量 指定该位置的平均流量 (1–999,999)。这些单位由以上选定的单位指定。 所需的时间量 指定执行样品程序期间要采集的总量 (10–10,000 ml)。 采集时长 指定总采集时长,以小时和分钟为单位 (0:01–999:00)。 最小样本量 如果流量过低,不足以达到最小样本量,则跳过此样品 (10–10,000 ml)。 切勿使用可变样品量 1. 为该选项选择或输入数值。 选项 说明 采集第一个样品 选择立即启动程序或在首次时间间隔过后再启动。 基于流量的采集 为基于流量的采样指定流量源: • 计数 - 请参阅 针对流量源使用计数 第 122。 • 4–20 mA 输入 - 请参阅 为流量源使用 4–20 mA 输入 第 122。 中文 121 针对流量源使用计数 1. 选择选项或为每个选项输入数值。 选项 说明 每次达到计数时采集样品 以固定的流量间隔采样,以计数为单位(1–9999 个计数)。 超时时间 如果流量很低,则采集样品。 时间 两次采样之间的最大间隔时间 (0:01–999:00)。每次采样之后,计时器都会重新启 动。 采集第一个样品 程序立即启动,或在第一个间隔时间之后启动。 为流量源使用 4–20 mA 输入 1. 选择选项或为每个选项输入数值。 选项 说明 选择单位 选择要使用的流量单位。 设定 4mA 输入 输入与 4mA 输入相一致的流量。 设定 20mA 输入 输入与 20mA 输入相一致的流量。 每次达到计数时采集样 品 输入抽样之前的总流量。 超时时间 在流量异常低时,选择启用该功能进行采样。 时间 两次采样之间的最大间隔时间 (0:01–999:00)。每次采样之后,计时器都会重新启 动。 采样量 输入每个样品的采集量。如选择采集瓶/样品模式,每只采集瓶内一种样品的采样量 为 (100–10,000 mL)。如果选择了可变样品量,则将跳过此选项。 恢复默认设置 恢复选项用于将所有的程序参数设为默认值。 1. 按下 MENU(菜单)。 2. 进入 PROGRAM SETUP(程序设置)>RESTORE(恢复)。 3. 选择 YES(是)。 启动或停止一个采样程序 利用 RUN/HALT PROGRAM(运行/中止程序)键可以启动或停止一个程序。可以设置多达 12 个用户 自定义的启动/停止时间/日期。更改程序或系统的设置之前,必须停止采样程序。 1. 按下 RUN/HALT PROGRAM(运行/中止程序)。 2. 选择一个选项。 选项 说明 Start(启动) 启动当前加载的程序。其状态将变为 RUNNING(正在运行)。 Halt(中止) 临时停止程序。其状态将变为 HALTED(已中止)。 Resume(恢复) 该程序将从中止点恢复运行。其状态将变为 RUNNING(正在运行)。 Start over(重新开始) 采样程序将从开始点启动。其状态将变为 RUNNING(正在运行)。 End program(结束程序) 结束当前正在运行的程序。其状态将变为 COMPLETE(已结束)。 122 中文 手动操作 使用手动操作功能采集抓取样品、移动分配臂或操作泵。详情请参阅该手册扩充版。 采样量校准 采样量可以通过液样传感器自动进行校准,或采用手动校准。使用液样传感器校准体积时,调整传感 器,精确测量所有程式化体积。如果禁用了液样传感器,则必须手动校准采样量、冲洗量及雨水量。 对于典型应用,可以利用自来水进行采样量校准。如果样品的组成明显不同于典型的水样,则应利用 所需采集的液体来校准采样器。校准之前必须停止所有程序。 校准液样传感器 液体传感器的精度随采样液体的类型而异。例如,严重浑浊样品的采样量不会与清澈样品的采样量一 样精确。液体传感器可以利用水样进行校准,以提高采样量的精度。 1. 将进样管放入水样中。 2. 按下 MENU(菜单)。 3. 进入 PROGRAM SETUP(程序设置)>MODIFY SELECTED(修改选择项)>LIQUID SENSOR (液体传感器)>ENABLE(启用),以确保启用液体传感器。 4. 进入 SYSTEM SETUP(系统设置)>LIQ SENSOR CAL(液体传感器校准)。 5. 选择 PERFORM CAL(进行校准)。 6. 选择 START(开始)。泵反向运行以清洗进样管。随后泵以前进方向运行。 7. 请查看进样管中的样品流,确保样品流经液体传感器。 8. 样品流经液体传感器并且无气泡时,按下 STOP(停止)。进样管会自动进行清洗。液体传感器结 束校准。 9. 选择 DONE(结束校准)将会退出,或选择 REPEAT(重复校准)再次进行校准。 恢复默认校准 液样传感器的采样量校准可以恢复为默认出厂设置。 1. 选择 PROGRAM SETUP(程序设置)>MODIFY SELECTED(更改所选项)>LIQUID SENSOR (液样传感器)>ENABLE(启用),以确保启用液样传感器。 2. 按下 VOLUME CALIBRATION(采样量校准),然后选择 RESET CAL(重置校准)。此时采样 量校准被设为默认设置。并会显示“数值已被重置为 0”信息。 使用液体传感器校准采样量 在启用液体传感器的情况下校准采样量,以便微调采样量。 1. 进入 PROGRAM SETUP(程序设置)>MODIFY SELECTED(修改选择项)>LIQUID SENSOR (液体传感器)>ENABLE(启用),以确保启用液体传感器。 2. 将进样管放入水样或自来水中。 注: 利用水样获得最佳精度。 3. 将泵出口侧的管子放入一个量筒中。 4. 按下 VOLUME CALIBRATION(采样量校准)并选择 CALIBRATION(校准)。 5. 选择 START(开始)。泵反向运行以清洗进样管。随后泵以前进方向运行,并采集采样程序中指 定的采样量。泵反向运行以清洗进样管。 6. 结束采样时,选择 DONE(结束)。将量筒中的采样量与程序设置菜单中输入的采样量进行对比。 7. 如采集体积与当前程序的采样量不同,输入实际采集的体积。按 Enter(输入)。传感器经过了调 整,以便精确地测量所有设置的采样量。 8. 选择 START(开始)以启动采样程序,或选择 CANCEL(取消)以退出采样量校准菜单。 中文 123 手动校准采样量 禁用液体传感器时,必须按时手动校准所有的采样量。采样量校准为当前程序规定的量。如果改变了 程序中的采样量,必须为新的采样量进行校准。 1. 选择 PROGRAM SETUP(程序设置)>MODIFY SELECTED(修改选择项)>LIQUID SENSOR (液体传感器)>DISABLE(禁用),以确保禁用液体传感器。 2. 将进样管放入水样或自来水中。 3. 将泵出口侧的管子放入一个量筒中。 4. 按下 VOLUME CALIBRATION(采样量校准)。如果显示多个采样量,则选择其中一个进行校 准。 5. 选择 START(开始)。泵反向运行以清洗进样管。随后泵以前进方向运行,并开始采集样品。 6. 到达所选体积或冲洗位置时,停止泵运行: • 采样量:获得主程序中指定的采样量时,按下 STOP(停止)。 • 冲洗:进样管中的液体到达液体检测器时,按下 STOP(停止)。 • 雨水采样量:获得雨水程序中指定的采样量时,按下 STOP(停止)。 7. 选择 DONE(结束校准) 进入下一个采样量校准,或是选择 REPEAT(重复校准)再次进行校 准。 8. 校准完所有采样量时,选择 DONE(结束校准)。 9. 将出样管连接到采样器的管接头上。 10. 选择 START(开始)启动采样程序,或选择 CANCEL(取消) 退出采样。 验证采样量 为确保采样量准确,需要进行随机采样。不要返回校准菜单检查体积,因为在校准开始时体积补偿已 被重置为 O。 1. 2. 3. 4. 5. 6. 7. 8. 按下 MANUAL OPERATION(手动操作)。 选择 GRAB SAMPLE(随机采样)。 将进样管放入水样中。 将泵出口侧的管子放入一个量筒中。 如果启用了液样传感器,则输入需要验证的采样量。 如果禁用了液样传感器,则输入程序中指定的采样量。 按下 Enter(输入)。随后泵循环将会启动(吹扫 - 采样 - 吹扫)。 将量筒中采集的样品量与预期样品量进行对比。如果采集的样品量不正确,则重新进行采样量校 准。 维护 危险 多种危险。只有合规的专业人员才能从事文档中本部分所述的任务。 危险 电击致命危险。执行维护或维修活动前,请断开仪器的电源连接。 警告 生物危害暴露危险。接触取样瓶和采样器部件时,请遵守安全处置规定。 124 中文 警告 多种危险。技术人员必须确保完成维护程序后,安全正确地操作设备。 清洁仪器 警告 火灾危险。切勿使用易燃试剂来清洁仪器。 注意 请勿使用任何液体清洁控制器室加热器。 如果水量不足以清洁控制器和泵,则断开控制器,然后使控制器远离采样器。在重新安装零件并恢复 使用之前,让控制器和泵充分干燥。 按如下方式清洁采样器: • 采样器外壳 - 使用湿布和温和的清洁剂清洁采样器外壳的内外表面。不得使用磨蚀性洗涤剂或溶 剂。 • 采样瓶 - 使用软刷、水和温和的清洁剂清洗取样瓶。先用清水冲洗,然后用蒸馏水冲洗。如有需 要,使用高压灭菌器清洁玻璃瓶。 更换干燥剂 控制器中的干燥剂筒用于吸收湿气,防止发生腐蚀。通过窗口 (图 12) 来监控干燥剂的颜色。新鲜的干 燥剂为黄色。颜色变绿时,需要更换干燥剂。 1. 2. 3. 4. 5. 6. 拧松后拆下干燥剂筒 (图 12)。 取下塞子,丢弃旧干燥剂。 将新鲜的干燥剂充满干燥剂管。 安装塞子。 为 O 形环涂上密封脂。 将干燥剂管安装在控制器中。 图 12 干燥剂筒 1 塞 3 O 形环 2 干燥剂管 4 干燥剂窗口 中文 125 泵维护 警告 夹伤危险。执行维护或维修活动前,请断开仪器的电源连接。 更换泵管 注意 除了制造商供应的管子之外,使用其他的管子会造成机械零件过度磨损和/或使泵性能不良。 检查泵管是否因为滚柱摩擦管子而产生磨损。显示提示信息时,或是管子出现磨损迹象时,请更换泵 管。 处理前必要条件: • 泵管 — 4.6 m 或 15.2 m(15 ft 或 50 ft)预切管材或成批管材 1. 2. 3. 4. 5. 断开控制器的电源。 如果使用成批管材,需要切割管材并标注对齐点。 参阅制造商网站上的扩展手册。 按照随后的图示步骤拆下泵管。 清洁泵壳内部和滚柱残留的硅树脂。 按照随后的图示安装新泵管。 126 中文 清洁转子 用柔性洗涤剂清洁转子、泵管路及泵外壳。请参阅 更换泵管 第 126 及随后图示步骤。 中文 127 更换分配臂管 在多瓶采样过程中,分配器分配臂会移动到每个采集瓶上。分配臂中的管子磨损时,需要进行更换。 确保在分配器和分配器分配臂中使用正确的管。 注: 分配器导管系统不同于泵的导管系统。装在分配器组件内的泵导管可能损坏分配器。而且,样品会因为分配臂 无法轻易移动而漏失。 1. 2. 3. 4. 从分配臂和中心部分顶部拆下管子。 将新管子插入分配臂中。确保导管符合 图 13 中规定的尺寸。 将管子的另一端插入中心部分顶部的接头中。 对分配器完成诊断测试,以确保运行正常。 图 13 分配器组件 1 管嘴 3 分配器电机 2 分配器分配臂 4 泵轴 5 分配器管 保险丝更换 保险丝不能由用户更换。请联系技术支持部门。保险丝熔断则意味着采样器出现问题,需要维修。 128 中文 目次 仕様 ページの 129 操作 ページの 142 総合情報 ページの 131 メンテナンス ページの 148 設置 ページの 134 より詳しい情報 より詳しい情報は、当社 Web サイトにて入手できます。 仕様 この仕様は予告なく変更されることがあります。 SD900 Portable Sampler 仕様 標準ベース コンパクトベース 混合ベース 寸法 直径: 50.5 cm 直径: 44.1 cm 直径: 50.28 cm 高さ: 69.4 cm 高さ: 61 cm 高さ: 79.75 cm 重量 15 kg、1 L ポリエチレンボトル (24x) を使 12.2 kg、575 mL ポリエチレン 用 ボトル (24x) を使用 14.8 kg 、10 L ポリエチレンコンテナー (1x) を使用 15 kg、950 mL ガラスボト ル (12x) を使用 12.9 kg、10 L ポリエチレンコン テナー (1x) を使用 筐体 衝撃耐性のある ABS、3 セクション構造、2.54 cm の二重壁ベース絶縁—ボトルが氷と直に接触 試料 温度 0 ~ 60 °C スト レー ナー 316 ステンレススチール (標準サイズ、高速、浅い水深用ロープロファイル、Teflon® 加工) または 316 ステンレススチール (標準サイズ) 試料 取入 管 9.5 mm 内径、ビニールまたはテフロン加工ポリエチレン 試料 ボト ルの 容量 1 L ポリエチレンおよび/または 350 mL ガラスボトル (24x) 575 mL ポリエチレンボトル (24x) 2.3 L ポリエチレンおよび/または 1.9 L ガラスボトル (8x) 950 mL ガラスボトル (8x) 3.8 L ポリエチレンおよび/または 3.8 L ガラスボトル (4x) 10 L ガラスボトル (1x) 21 L ポリエチレンボトル (1x) 10 L ポリエチレンボトル (1x) 3.8 L ポリエチレンおよび/または 3.8 L ガラスボトル (2x) 21 L もしくは 15 L ポリエチレン複合材 コンテナー、または 20 L ポリエチレン、 10 L ポリエチレン、もしくは 10 L ガラス (1x) SD コントローラー 仕様 詳細 寸法 (W x H x D) 29.2 x 17.1 x 26.4 cm 重量 4.2 kg 日本語 129 仕様 詳細 筐体 PC/ABS ブレンド、NEMA 4X、6、IP 67 電源 8754500 電源から 15 VDC が供給。 内蔵電源から 15 VDC が供給 過負荷保護 ポンプ用の 7 A、DC 回路のヒューズ 動作周囲温度 0 ~ 50 °C 保管温度 –30 ~ 60 °C 保管/作動湿度 100 %、結露可 ポンプ スプリング式ナイラトロンローラーによる高速蠕動ポンプ ポンプ筐体 IP37 ポンプチューブ 内径 9.5 mm x 外径 15.9 mm (内径 3/8 インチ x 外径 5/8 インチ) シリコン ポンプチューブの寿命 20,000 サンプルサイクル: サンプル量 1 L、洗浄回数 1 回、ペース間隔 6 分、長さ 4.9 m の 3/8 インチ取入管、揚水高さ 4.6 m、サンプル温度 21 °C の場合 サンプル揚水高さ 最大 8.5 m: 長さ 8.8 m の 3/8 インチビニール製取入管、海抜、温度 20 ~ 25 °C の場 合 ポンプ流量 4.8 L/min (1.25 gpm): 揚水高さ 1 m、3/8 インチ取入管の場合 サンプル量 10 ~ 10,000 mL の範囲で 10 mL 単位で設定可能 サンプル量の繰り返し精 度 (標準値 サンプル量 200 mL で ±5 %: 未校正の液体、揚水高さ 4.6 m、長さ 4.9 m の 3/8 イン チビニール製取入管、単一ボトル、フルボトルシャットオフ、室温、高度 1524 m の 場合 サンプル量の精度 (標準 値) サンプル量 200 mL で ±10 %: 未校正の液体、揚水高さ 4.6 m、長さ 4.9 m の 3/8 イン チビニール製取入管、単一ボトル、フルボトルシャットオフ、室温、高度 1524 m の 場合 サンプリングモード ペース: 時間固定、流量固定、時間可変、流量可変、一定時間可変量 (CTVV)。 このマニュアルの拡張版の一定時間可変量 (CTVV) サンプリングを参照してくださ い。 分配: 単一ボトル複合、複数ボトル複合、複数ボトル分離、サンプルあたりのボトル 数、ボトルあたりのサンプル数、サンプルあたりのボトルの組み合わせ、ボトルあた りのサンプル数 実行モード ユーザーが入力したサンプル数で連続および非連続 保存可能なプログラム数 最大 3 つのサンプリングプログラムを保存可能 移送速度 (標準値) 0.9 m/s: 揚水高さ 4.6 m、長さ 4.9 m の 3/8 インチビニール製取入管、21 °C、高度 1524 m の場合 液体センサ 超音波. 本体: ウルテム NSF ANSI 規格 51 認定、USP Class VI 準拠 サンプルの履歴 最大 510 レコード エアーパージ 各サンプルの前後で自動的にエアーパージ。パージ時間は取入管の長さに応じて自 動補正。 イベントログ 510 件 接続 電源、補助、シリアル通信、ディストリビューター、SDI-12 接液面の材質 ステンレススチール、ポリエチレン、テフロン、ウルテム、シリコン 通信 RS232、Modbus、SDI-12 保証 1年 130 日本語 ® 総合情報 いかなる場合も、製造元は、例えそのような損害が生じる可能性について報告を受けていたとして も、本マニュアルに含まれるいかなる瑕疵または脱落から生じる直接的、間接的、特定、付随的また は結果的に生じる損害に関して責を負いません。製造元は、通知または義務なしに、随時本マニュア ルおよび製品において、その記載を変更する権利を留保します。改訂版は、製造元の Web サイト上 にあります。 安全情報 告知 メーカーは、本製品の目的外使用または誤用に起因する直接損害、偶発的損害、結果的損害を含むあらゆる損害 に対して、適用法で認められている範囲で一切責任を負わないものとします。ユーザーは、適用に伴う危険性を 特定したり、装置が誤作動した場合にプロセスを保護するための適切な機構を設けることに関して、全責任を負 うものとします。 この機器の開梱、設定または操作を行う前に、このマニュアルをすべてよく読んでください。危険お よび注意の注意事項に注意を払ってください。これを怠ると、オペレータが重傷を負う可能性、ある いは機器が損傷を受ける可能性があります。 本装置に備わっている保護機能が故障していないことを確認します。本マニュアルで指定されてい る以外の方法で本装置を使用または設置しないでください。 危険情報の使用 危険 回避しなければ死亡または重傷につながる、潜在的または切迫した危険な状況を示します。 警告 避けない場合、死亡事故や負傷が起こるかも知れない危険な状況を示します。 注意 軽傷または中傷事故の原因となる可能性のある危険な状況を示しています。 告知 回避しなければ、装置の損傷を引き起こす可能性のある状況を示します。 特に注意を要する情報。 使用上の注意ラベル 測定器上に貼付されたラベルや注意書きを全てお読みください。これを怠ると、人身傷害や装置の損 傷につながるおそれがあります。測定器に記載されたシンボルは、使用上の注意と共にマニュアルを 参照してください。 これは安全警報シンボルです。潜在的な障害を避けるためにこのシンボルのすべて安全メッセージに 従ってください。 装置上では、作業または安全情報に関しては取り扱い説明書を参照してください。 このシンボルは、指や皮膚を挟み込む可能性があることを示します。 このシンボルが付いている電気機器は、 ヨーロッパ域内または公共の廃棄処理システムで処分できま せん。古くなったり耐用年数を経た機器は、廃棄するためにメーカーに無償返却してください。 取得認証 カナダの障害発生機器規則、IECS-003、クラス A: 日本語 131 テスト記録のサポートはメーカーにあります。 このクラス A デジタル装置はカナダの障害発生機器規則の要件をすべて満たします。 Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences. FCC PART 15、クラス 「A」 限度値 テスト記録のサポートはメーカーにあります。この機器は FCC 規則のパート 15 に準拠します。運 転は以下の条件を前提としています: 1. この装置が有害な干渉の原因とならないこと。 2. この装置が望ましくない動作の原因となる可能性のあるいかなる干渉にも対応しなければなりま せん。 これらの規格への準拠に責任を持つ当事者による明示的承認を伴わなずにこの装置に対する改変ま たは改造を行うと、ユーザーはこの機器を使用する権限を失う可能性があります。この装置は、FCC 規則のパート 15 に従って、クラス A のデジタル機器の制限に準拠することが試験によって確認され ています。これらの制限は、この機器が商用の環境で使用されたときに、有害な干渉から適切に保護 することを目的に設定されています。この機器は、無線周波数エネルギーを生成および使用するもの で、取り扱い説明書に従って取り付けおよび使用しない場合にはそれを放射する場合があり、無線通 信に対して有害な干渉を発生させる可能性があります。住宅地域における本装置の使用は有害な電 波妨害を引き起こすことがあり、その場合ユーザーは自己負担で電波妨害の問題を解決する必要があ ります。以下の手法が干渉の問題を軽減するために使用可能です。 1. 装置から電源を取り外して、電源が干渉源かどうかを確認します。 2. 装置が干渉を受けている装置と同じコンセントに接続されている場合は、装置を別のコンセント に接続してください。 3. 妨害を受けている装置から本装置を離します。 4. 干渉を受けるデバイスの受信アンテナの位置を変更します。 5. 上記の手法を組み合わせてみます。 製品の概要 危険 化学的および生物学的な危険。この装置の用途が処理工程や薬液注入システムの監視であり、それらに 対して公衆衛生、公衆安全、食品/飲料の製造/加工に関する規制や監視要件が存在する場合、この装置 の使用者には、該当するすべての規制を把握して遵守する責任、および装置の異常時に関する当該規制 に従って十分かつ適切な措置を講じる責任があります。 注意 火災の危険。本製品は、可燃性の液体を使用するように設計されていません。 SD900 Portable Sampler は液体試料を指定間隔で収集して、ボトルまたはリテイナーに保管します。 サンプラーはさまざまな水性用途、および有毒汚染物質や浮遊固体に使用します。さまざまなリテイ ナー、ボトル、またはコンテナーを使用してサンプラーをセットアップします。図 1 を参照してくだ さい。 132 日本語 図 1 製品の概要 1 コンパクトベース 6 コントローラーの接続部 2 標準絶縁ベース 7 ポンプ 3 センターセクション 8 コントローラー 4 電源 9 液体センサ 5 上部カバー 10 標準絶縁ベース、21 L コンテナー用 製品コンポーネント すべてのコンポーネントを受け取っているか確認します。図 2 を参照してください。品目が不足し ていたり損傷していたりした場合は、直ちに HACH Japan または弊社販売代理店にお問い合わせく ださい。 図 2 製品コンポーネント 1 ベース (標準、コンパクト、または混合) 6 バッテリー充電器 (オプション) 2 単一のボトル使用する場合のコンポーネント (ボト ルおよびサポートは変更可能) 7 AC 電源 (オプション) 3 コントローラーを含むセンターセクション 8 ストレーナー 4 バッテリー (オプション) 9 取入管 (ビニールまたはテフロン加工) 5 上部カバー 10 複数のボトル使用する場合のコンポーネント (ボ トルおよびリテイナーは変更可能) 日本語 133 設置 危険 複合的な危険。本書のこのセクションに記載されている作業は、必ず資格のある要員が行う必要があり ます。 閉鎖空間に関する注意事項 危険 爆発の危険。プリエントリ試験トレーニング、換気、エントリ手順、排気/レスキュ手順および安全作業 の実行は閉じ込められたスペースに入る前に必要です。 次に示す情報は、閉鎖空間への立ち入りに関連して発生する危険およびリスクについて理解するため のものです。 1993 年 4 月 15 日に、CFR 1910.146「Permit Required Confined Spaces」に関する OSHA の最終 的な決定が法制化されました。この標準は米国の 250,000 を超える産業用地に直接影響し、閉鎖空間 で働く作業者の健康と安全を保護することを目的としています。 閉鎖空間の定義: 閉鎖空間は、次の条件が 1 つ以上該当する (またはすぐにでも該当する可能性のある) 場所または密 閉状態のことです。 • 酸素濃度が 19.5 % 未満また 23.5 % を超える、あるいは硫化水素 (H2S) 濃度が 10 ppm を超える 雰囲気 • ガス、蒸気、霧、塵、または繊維による引火性のある雰囲気、または爆発の可能性がある • 接触または吸引すると怪我、健康障害、または死亡を引き起こす可能性のある有毒物質が存在する 閉鎖空間は人間が滞在することを意図していません。閉鎖空間への立ち入りが制限され、既知の危険 が存在するか、危険が発生する可能性があります。閉鎖空間の例には、マンホール、煙突、パイプ、 タンク、スイッチ保管所、およびその他の同様な場所などです。 危険なガス、蒸気、霧、塵、または繊維が存在する可能性のある閉鎖的な空間または場所に立ち入る 場合は、事前に標準的な保護手順を実行する必要があります。閉鎖空間に立ち入る前に、閉鎖空間へ の立ち入りに関連したすべての手順を検索し、参照してください。 設置 現場設置ガイドライン 危険 爆発の危険。この装置は、危険がある場所での設置は許可されていません。 設置現場の評価については、以下のガイドラインを参照してください。 • サンプラーを密閉空間に取り付ける場合は、すべての安全注意事項に従ってください。閉鎖空間に 関する注意事項 ページの 134 を参照してください。 • 設置場所の温度が指定範囲内であるか確認してください。仕様 ページの 129 を参照してくださ い。 • サンプラーは水平面に取り付けるか、サスペンションハーネス、サポートブラケット、またはスパ ナーバーを使用してサンプラー吊り下げてください。サンプラーをマンホールに取り付け ページの 135 および該当する設置説明書を参照してください。 • 分析遅延を小さくするために、可能な限り試料源の近くに配置してください。サンプラーの配管 ページの 137 を参照してください。 • 輸送速度および最大揚水高さの制限については、仕様 ページの 129 を参照してください。 134 日本語 サンプラーをマンホールに取り付け マンホールの試料水の上にサンプラーを取り付けます。サンプラーを取り付けるには、スパナーバー またはサポートブラケットを使用します。マンホールの内側にスパナーバーを取り付けます。スパ ナーバーは壁に対する圧力で支えます。サポートブラケットはマンホールカバーと同じ幅です。サ ポートブラケットはカバーの真下に取り付けて支えます。メーカーの Web サイトにある拡張版マニ ュアルを参照してください。サンプラーの取り付け手順については、アクセサリーに付属している説 明書を参照してください。 ディストリビューターまたはフルボトルシャットオフの取り付け (オプション) ディストリビューターまたはフルボトルシャットオフアセンブリーは通常、工場出荷時に取り付けら れています。取り付け手順については、ディストリビューターまたはフルボトルシャットオフの説明 書を参照してください。 サンプラーの準備 サンプルボトルの洗浄 ブラシ、水、および低刺激の洗剤を使用して、サンプルボトルとキャップを洗浄します。コンテナを 真水で洗浄してから、蒸留水ですすぎます。 単一のボトルの取り付け 混合試料を 1 つ収集するには、単一のボトルを使用します。ボトルがいっぱいになると、フルボトル シャットオフにより試料プログラムが停止します。 1. サンプルボトルを洗浄します。サンプルボトルの洗浄 ページの 135 を参照してください。 2. 図 3 に示しているようにサンプルボトルを取り付けます。 図 3 単一ボトルの取り付け 1 ポリエチレンボトル、10 L (1918) 6 ポリエチレンボトル、21 L (6494) 2 ガラスボトル、10 L (6559) 7 コンパクトベース (8975) 3 サポート (1502) 8 標準絶縁ベース (8976) 4 ポリエチレンボトル、15 L (1367) 9 標準絶縁ベース、21 L ボトル用 (8561) 5 ポリエチレンボトル、19 L (6498) 最初のボトルの取り付け (複数の試料を収集する場合) 複数の試料を個別のボトルまたは 2 本以上のボトルに収集する場合は、複数のボトルを使用します。 ディストリビューターで試料チューブを各ボトルの上に移動します。図 4 に示すように、サンプラー 日本語 135 ベースにボトルを取り付けます。サンプラーベースのラベルの下に、最初の試料ボトル (番号 1) を取 り付けます。残りのボトルをラベルが示す方向に昇順に取り付けます。必要なコンポーネントの図 については、図 5 ページの 137 を参照してください。 図 4 ボトル番号 1 の取り付け 1 ボトル番号 1 の場所 (コンパクトベース) 6 ゴム製ストラップ 2 ボトル番号 1 の場所 (ボトル数 24) 7 標準絶縁ベース、21 L ボトル用 3 ボトル番号 1 の場所 (ボトル数 8) 8 標準絶縁ベース 4 ボトル番号 2 の場所 (ボトル数 2 または 4) 9 コンパクトベース 5 リテイナー 複数のボトルの取り付け 複数のボトルを取り付ける場合は、ディストリビューターアームで試料チューブを各ボトルの上に移 動します。指定した数の試料が収集されると、試料の収集が自動的に停止します。 1. サンプルボトルを洗浄します。サンプルボトルの洗浄 ページの 135 を参照してください。 2. サンプラーベースのラベルの下に、最初の試料ボトル (番号 1) を取り付けます。最初のボトルの 取り付け (複数の試料を収集する場合) ページの 135 を参照してください。 3. 図 5 に示しているようにサンプルボトルを組み立てます。ボトル数が 8 本以上の場合は、最初の ボトルをボトル 1 インジケーターの横 (時計回り方向) に配置します。 4. ボトルアセンブリーをサンプラーに配置します。ボトル数が 8 本以上の場合は、下部トレイの溝 にワイヤを揃えます。 136 日本語 図 5 複数のボトルの取り付け 1 ポジショナー/リテイナー (2347) 8 ガラスボトル、3.8 L (4x) (2216) またはポリエチレ ンボトル、3.8 L (4x) (2217) 2 リテイナー (1422) 9 ガラスボトル、1.9 L (8x) (1118) またはポリエチレ ンボトル、2.3 L (8x) (657) 3 リテイナー (2190) 10 ポリエチレンボトル、1 L (24x) (737) 4 リテイナー (2189) 11 ガラスボトル、350 mL (24x) (732) 5 ガラスボトル、950 mL (8x) (2348) 12 コンパクトベース (8975) 6 ポリエチレンボトル、575 mL (24x) (1369) 13 標準ベース (8976) 7 ガラスボトル、3.8 L (2x) (2214) またはポリエチレ ンボトル、3.8 L (2x) (2215) サンプラーの配管 ストレーナーを (表面付近や下部ではなく) 試料水流の中央に取り付けて、代表的な試料が収集される ようにします。図 6 を参照してください。取入管の取り付け手順については、図 7 を参照してくださ い。 1. 図 7 に示しているようにチューブをサンプラーに接続します。 注: テフロンライニングチューブを使用する場合は接続キット 2186 を使用します。 2. 取入管とストレーナーを試料源の本流に設置し、水が攪拌されて十分に混合されるようにします。 • 取入管はできるだけ短くしてください。 • サンプルを採取するたびにチューブが完全に排水されるよう、取入管の傾斜が保たれるように してください。 注: 傾斜を保つことができない場合、またはチューブが加圧される場合は、液体センサを無効にしてくださ い。試料量を手動で調整します。サンプル量の手動校正 ページの 147 を参照してください。 • 取入管がつぶれていないことを確認してください。 日本語 137 図 6 器具の取り付け 1 ストレーナー 3 取入管 2 揚水高さ 図 7 取入管の取り付け 配線 コントローラーの接続部 警告 電気ショックの危険。外部接続された機器には、該当する国の安全標準評価が必要です。 図 8 に、コントローラーの接続部を示します。 138 日本語 図 8 コントローラーのコネクタ 1 補助デバイス 4 SDI-12 デバイスオプション 2 電源 5 ディストリビューター/フルボトルシャットオフ 3 シリアル通信 流量計の接続 サンプルの流量が指定値を上回ったか下回った場合にサンプラーを開始または停止させるには、コン トローラーに流量計を接続します。 用意するもの: • Sigma 流量計用の多目的フルケーブル (またはモデル 980 流量計用の 980 ハーフケーブル) • 追加の接続用のオプションのスプリッター。複数のスプリッターを直列接続できます。 1. ケーブルの一方の端を流量計に接続します。モデル 980 流量計の場合は、モデル 980 流量計の取 扱説明書を参照してください。 2. ケーブルのもう一方の端をコントローラーの補助デバイスコネクタに接続します。 注: 流量計のケーブルが 6 ピンの場合は、6 ピン -> 7 ピン変換アダプターケーブルを使用してください。 PC または通信ネットワークの接続 データを転送したりサンプラーを設定したりするには、PC または通信ネットワークをコントローラ ーに接続します。 用意するもの: • シリアルケーブル、7 ピン RS232 -> DB-9 (図 9 を参照)。接続: B -> 5 (信号接地)、D -> 3 (RCD)、 F -> 2 (TXD)、G (接地)。 1. シリアルケーブルの一方の端を通信デバイスまたはネットワークに接続します。 2. ケーブルのもう一方の端をコントローラーのシリアル通信コネクタに接続します。 図 9 シリアル通信ケーブル 日本語 139 サンプラーを電源に接続 危険 感電死の危険。この装置を屋外または湿っている可能性がある場所で使用する場合は、主電源との接続 に漏電回路安全装置 (GFCI/GFI)を使用する必要があります。 危険 火災の危険。電力線に 15 A の回路ブレーカーを取り付けます。回路ブレーカーは、機器の近くにあれ ば、ローカルの電源切断装置になります。 危険 感電死の危険。保護アース接地(PE)接続が必要です。 警告 感電死の危険。ローカル電源切断へのアクセスが容易であることを確認します。 警告 電気ショックの危険。電源はポンプサイクル間隔が短すぎる場合に過熱することがあります。 告知 ポンプサイクル時間によって電源が過熱状態にならないことを確認します。ポンプのサイクル時間の計算 ページの 141 を参照してください。 サンプラーをバッテリーまたは AC 電源に接続します。 下図に示した手順を参照してください。 140 日本語 ポンプのサイクル時間の計算 電源が過熱しないように、ポンプはサンプルサイクル間で一定時間以上オフになっている必要があり ます。 1. サンプルサイクル中のポンプの合計連続運転時間を計算します。計算には、プレパージ、取入管 洗浄、サンプル、サンプル採取リトライ、ポストパージというすべての段階を含めます。 2. 図 10 を参照して、サンプルサイクル間でポンプがオフになっていなければならない最小時間を見 つけます。 3. サンプルサイクル間でポンプがこの最小時間にわたってオフになるように、サンプラーのプログ ラムのペースまたは時間間隔が設定されていることを確認します。 図 10 ポンプのオン/オフ時間 (50 °C 時) 1 ポンプがオフになっていなければならない時間 (分 単位) 3 例: ポンプが 5 分間連続的にオンになる場合、ポン プは 15 分間オフになる必要があります。 2 ポンプが連続的にオンになっている時間 (分単位) 電源の投入 電源ボタンを押すと、電源のオン/オフを切り替えることができます。電源 (AC 電源またはバッテリ ー電源) が正しく取り付けられていることを確認してください。 ユーザーインターフェースとナビゲーション コントローラーのキーパッドを図 11 に示します。[MENU (メニュー)] キーは、サンプリングプログ ラムの設定、コントローラーの設定、診断テストの実行に使用します。矢印キー、[ENTER (入力)] キ ー、および [BACK (戻る)] キーは、メニューのスクロール、項目の選択、値の入力に使用します。デ ィスプレイ画面の下または横に矢印が表示されている場合は、選択可能なオプションがさらにあるこ とを示しています。 日本語 141 図 11 コントローラーのキーパッド 1 POWER (電源) 5 MENU (メニュー) 9 STATUS (状態) 2 VOLUME CALIBRATION (量の 校正) 6 ENTER (入力) 10 RUN/HALT PROGRAM (プロ グラムの実行/中止) 3 STOP (停止) 7 BACK (戻る) 4 LED 8 矢印キー 11 MANUAL OPERATION (手動操 作) 操作 システム設定の指定 1. [MENU (メニュー)] を押します。 2. [SYSTEM SETUP (システム設定)] に進みます。 3. 該当するオプションを変更します。 オプション 説明 Time/date setup (日時設定) 時刻 (24 時間形式) と日付を設定します。 Communication (通信) シリアルポートのボーレート (19200、38400、57600、または 115200) とプロ トコル (Modbus RTU または ASCII) を選択します。 Setup base (ベースの設定) サンプラーベースを選択します (全天候型冷蔵式、冷蔵式、携帯型標準、また は携帯型コンパクト)。 Language(言語) コントローラーの言語を選択します。 Liq sensor cal (液体センサ 校正) 液体センサを校正するか、デフォルトに設定します。 Set contrast (コントラスト 設定) LCD 画面のコントラストを調整します。 Pump duty cycle (ポンプの デューティサイクル) ポンプのデューティサイクルを変更します。範囲: 50 ~ 100 % (デフォルトは 100 %)。 142 日本語 オプション 説明 Password setup (パスワー ド設定) プログラム設定メニューおよびシステム設定メニューへのアクセスを制御す るためのパスワードを設定します。[ENABLE (有効)] > [YES (はい)] を選択し て新しいパスワードを設定するか、[ENABLE (有効)] > [NO (いいえ)] を選択し て現在のパスワードを有効にします。パスワードを初めて設定する場合は、現 在のパスワードに 900900 を使用してください。 Tubing life (チューブ寿命) ポンプチューブの変更が必要になったときにそれを知らせる通知機能を設定 します。ポンプのサイクルカウントを 0 にリセットするには、[ENABLE (有 効)] > [RESET CYCLES (サイクルのリセット)] に進みます。通知機能に使用 するポンプのサイクル数を変更するには、[ENABLE (有効)] > [CYCLE LIMIT (サイクル上限)] に進みます。ポンプのサイクルカウントの現在の値を表示す るには、[ENABLE (有効)] > [STATUS (ステータス)] に進みます。 SDI–12 (オプション) SDI–12 デバイスの動作を設定します。このオプションはデバイスが検出され た場合のみ表示されます。 サンプリングプログラムの設定 サンプリングプログラムには、サンプルを自動的に採取するのに必要なすべてのパラメーターが含ま れています。サンプルは一定の間隔で、またはサンプリングプログラムが完了したときに採取されま す。 1. [MENU (メニュー)] を押します。 2. [PROGRAM SETUP (プログラム設定)] > [MODIFY ALL (すべて修正)] に進みます。最初の画面が 表示されます。 3. 各パラメーターの値を選択または入力します。左または右へ移動するには、左または右矢印キー を使用します。値を変更するには上または下矢印キーを使用します。矢印キーを使用して、必要 に応じて単位 (gal または mL) を変更します。 オプション 説明 Bottles (ボトル) ボトル数量: サンプラーに装填されているボトルの数 (1、2、4、8、または 24)。 Full BTL shutoff (フルボトルシャットオフ): Enabled (有効)/Disabled (無効)。有効に した場合、コントローラーはボトルのオーバーフロー状態を確認します。 Bottle volume (ボトル容量): 各ボトルの容量。範囲: 50 ~ 65,000 mL Intake tubing (取入管) Tubing length (チューブの長さ): ストレーナーから液体センサまでの取入管の長さ。 範囲: 100 ~ 3000 cm。正確なサンプル量を得るためには正確な長さを入力する必 要があります。 Intake tube type (取入管のタイプ): 取入管の径と材質 (1/4 インチビニール、3/8 イン チビニール、または 3/8 インチテフロン)。 Program delay (プロ グラム遅延) Sample pacing/collection (サ ンプルのペース/採取) Enabled (有効)/Disabled (無効): 有効にすると、サンプリングプログラムは指定され た時間またはカウント数に達したときに開始されます。Date and time (日時): プロ グラムが開始する日時 (24 時間形式)。Counts (カウント): プログラムが開始する、 フローメーターからのカウント数またはパルス数 (1 ~ 9999 カウント)。サンプル のペースを時間に設定した場合、プログラム遅延は無効になります。 時間ベース - 時間ベースの採取 ページの 144 を参照してください。 注: CTVV を使用する場合: • • • • • 液体センサは必ず有効にします。 サンプルの分配は複合です。 実行モードは非連続です。 最初のサンプルは時間間隔の後に採取されます。 フローソースは常に 4 ~ 20 mA 入力です。 流量ベース - 流量ベースの採取 ページの 145 を参照してください。 日本語 143 オプション 説明 Sample distribution (サンプルの分配) Deliver samples to all (すべてにサンプルを供給) > YES (はい) 各サンプルがすべてのボトルに供給されます。End after last (最後のサンプルの後 に終了): プログラムは最後のサンプルの後に終了します (1 ~ 999)。Continuous (連 続): プログラムは手動で停止するまで続行されます。 Deliver samples to all (すべてにサンプルを供給) > NO (いいえ) 試料は一部のボトルに供給されます 。Samples/bottle (サンプル/ボトル): 各ボトル に採取するサンプルの数 (1 ~ 999)。Bottles/sample (ボトル/サンプル): 同じサンプ ルが含まれるボトルの数。 Liquid sensor (液体セ ンサ) 有効または無効。無効の場合、サンプル量は時間で校正する必要があります。 Sample volume (サン プル量) 各サンプルの量 (mL 単位)。ボトル/サンプルモードを使用した場合、指定した量の サンプルが各ボトルに入れられます (10 ~ 10,000 mL)。サンプルの量がボトルの容 量よりも多くならないようにしてください。サンプルの量は 10 mL 単位で丸められ ます。 Intake rinses (取入管 洗浄) サンプルが採取される前に取入管が洗浄される回数です (0 ~ 3)。 Sample retries (サン プル採取リトライ) サンプリングが失敗した後に、サンプリングを再試行する回数です (0 ~ 3)。 Site ID (サイト ID) サンプリング場所の名前です (最大 12 文字)。プログラムを保存した場合、サイト ID はプリセット名として使用されます。 Advanced sampling (高度なサンプリング) 高度なサンプリングメニューに進みます。必要に応じて、[Advanced Sampling (高度 なサンプリング)] メニューのオプションを設定します。 Run program (プログ ラム実行) 基本的なサンプリングプログラムの作成を終了します。プログラムの開始またはキ ャンセルを求めるメッセージがサンプラーに表示されます。 時間ベースの採取 1. 各オプションの値を選択または入力します。 オプション 説明 Pacing interval (ペース 間隔) サンプルを一定の時間間隔で採取します。時間と分 (0:01 ~ 999:00) を指定しま す。 Take first sample (最初 プログラムは直ちに開始されるか、または最初の時間間隔後に開始されます。 のサンプルを採取) Variable volume (可変 採水量) サンプル量を変化させるかどうかを指定します (流量ベース)。変化させる場合は、 可変量を使用する ページの 144 を参照してください。変化させない場合は、可変 量を使用しない ページの 145 を参照してください。 可変量を使用する サンプルペースは一定時間可変量 (CTVV) です。 1. 各オプションの値を選択または入力します。 オプション 説明 単位選択 使用する流量単位を選択します。 Map 4 mA input (4 mA 入力マッ ピング) 4 mA 入力に相当する流量を入力します。 Map 20 mA input (20 mA 入力マ ッピング) 20 mA 入力に相当する流量を入力します。 144 日本語 オプション 説明 Average flow rate (平均流量) サイトの平均流量を指定します (1 ~ 999,999)。単位は前述の手順で選 択した単位です。 Time volume desired (必要な時 間量) サンプルプログラム中に採取される合計量を指定します (10 ~ 10,000 ml)。 Collection period (採取期間) 合計採取期間を時間および分単位で指定します (0:01 ~ 999:00)。 最小サンプル量 この最小量を採取できる流量がない場合、サンプルはスキップされます (10 ~ 10,000 ml)。 可変量を使用しない 1. オプションの値を選択または入力します。 オプション 説明 最初のサンプルを採取します。 プログラムが直ちに開始されるか、または最初の時間間隔後に開始されるか を選択します。 流量ベースの採取 流量ベースのサンプリングのフローソースを指定します。 • カウント - フローソースのカウントを使用する ページの 145 を参照してください。 • 4 ~ 20 mA 入力 - フローソースの 4 ~ 20 mA 入力を使用します。 ページの 145 を参照してくだ さい。 フローソースのカウントを使用する 1. 各オプションの値を選択または入力します。 オプション 説明 Take sample every (サンプル採 取間隔) 一定の流量間隔でサンプルを採取します。カウント (1 ~ 9999) を指定し ます。 Overrride time (サンプル採取間 隔) 流量が非常に少ない場合にサンプルを採取します。 Time (時間) サンプル間の最大時間 (0:01 ~ 999:00)。各サンプルが採取された後に、 タイマーが再度開始されます。 Take first sample (最初のサンプ プログラムは直ちに開始されるか、または最初の時間間隔後に開始されま ルを採取) す。 フローソースの 4 ~ 20 mA 入力を使用します。 1. 各オプションの値を選択または入力します。 オプション 説明 単位選択 使用する流量単位を選択します。 Map 4 mA input (4 mA 入力マッピング) 4 mA 入力に相当する流量を入力します。 Map 20 mA input (20 mA 入力マッピング) 20 mA 入力に相当する流量を入力します。 Take sample every (サン サンプルを採取した後の合計流量を入力します。 プル採取間隔) 割り込み時間 選択すると、流量が極端に低い場合に採取するサンプルの押し出しが有効になり ます。 日本語 145 オプション 説明 Time (時間) サンプル間の最大時間 (0:01 ~ 999:00)。各サンプルが採取された後に、タイマー が再度開始されます。 サンプル量 サンプルごとに採取する量を入力します。ボトル/サンプルモードを選択した場 合、指定した量のサンプルが各ボトルに入れられます (100 ~ 10,000 mL)。可変 量を選択した場合、このオプションはスキップされます。 デフォルト設定の復元 復元オプションを実行すると、すべてのプログラムパラメーターがデフォルト設定に設定されます。 1. [MENU (メニュー)] を押します。 2. [PROGRAM SETUP (プログラム設定)] > [RESTORE (復元)] に進みます。 3. [YES (はい)] を選択します。 プログラムの起動/停止 [RUN/HALT PROGRAM (プログラムの実行/中止)] キーを使用して、プログラムを起動または停止し ます。最大 12 個のユーザー定義の起動/停止日時を設定できます。プログラムまたはシステムの設 定を変更するには、プログラムは停止する必要があります。 1. [RUN/HALT PROGRAM (プログラムの実行/中止)] を押します。 2. いずれかのオプションを選択します。 オプション 説明 開始 現在ロードされているプログラムが開始します。ステータスが [RUNNING (実行中)] に変わります。 Halt (中止) プログラムが一時的に停止します。ステータスが [HALTED (中止)] に変わります。 Resume (再開) 中止された位置からプログラムが再開されます。ステータスが [RUNNING (実行中)] に変わります。 Start over (やり直し) プログラムが最初から開始されます。ステータスが [RUNNING (実行中)] に変わり ます。 End program (プログ ラム終了) 現在実行中のプログラムが停止します。ステータスが [COMPLETE (完了)] に変わり ます。 手動操作 グラブサンプルを採取したり、ディストリビューターアームを移動したり、ポンプを作動させたりす るには、手動操作を使用します。追加情報はこのマニュアルの拡張バージョンを参照してください。 量の校正 サンプル量は、液体センサを使用して自動的に校正することも、手動で校正することもできます。液 体センサを使用して量の校正を行った場合、プログラムされているすべての量を正確に測定するよう にセンサが調整されます。液体センサが無効になっている場合は、サンプル、洗浄、および雨水の量 を手動で校正する必要があります。 標準的な用途の場合は、量の校正に水道水を使用することができます。サンプルの組成が標準的な水 サンプルと大幅に異なる場合は、採取する液体でサンプラーを校正します。校正を実行する前に、す べてのプログラムを停止する必要があります。 液体センサの校正 液体センサの精度は、サンプリングされた液体の種類によって異なります。たとえば、濁度の高いサ ンプルでは、清浄な水の場合よりも量の精度が低くなることがあります。サンプル水で液体センサを 校正して、量の精度を向上させることができます。 146 日本語 1. 取入管をサンプル水に入れます。 2. [MENU (メニュー)] を押します。 3. [PROGRAM SETUP (プログラムの設定)] > [MODIFY SELECTED (選択修正)] > [LIQUID SENSOR (液体センサ)] > [ENABLE (有効)] に進み、液体センサが有効になっていることを確認し ます。 4. [SYSTEM SETUP (システム設定)] > [LIQ SENSOR CAL (液体センサ校正)] に進みます。 5. [PERFORM CAL (校正実行)] を選択します。 6. [START (開始)] を選択します。ポンプが逆転して、取入管がパージされます。その後、ポンプは 正転します。 7. 取入管のサンプルの流れを見て、サンプルが液体センサ内を流れていることを確認します。 8. サンプルが液体センサ内を流れ、気泡が存在しなくなったら、[STOP (停止)] を押します。取入管 が自動的にパージされます。これで液体センサの校正は完了です。 9. [DONE (完了)] を選択して終了するか、[REPEAT (繰り返し)] を選択して再度校正を実行します。 デフォルト校正の復元 液体センサの量の校正は、デフォルトの工場出荷設定に戻すことができます。 1. [PROGRAM SETUP (プログラム設定)] > [MODIFY SELECTED (選択修正)] > [LIQUID SENSOR (液体センサ)] > [ENABLE (有効)] を選択して、液体センサが有効になっていることを確認します。 2. [VOLUME CALIBRATION (量の校正)] を押して、[RESET CAL (校正リセット)] を選択します。量 の校正がデフォルト設定に設定されます。 「value has been reset to 0 (値が 0 にリセットされまし た)」というメッセージが表示されます。 液体センサを使用したサンプル量の校正 サンプル量を微調整するには、液体センサが有効な状態でサンプル量を校正します。 1. [PROGRAM SETUP (プログラムの設定)] > [MODIFY SELECTED (選択修正)] > [LIQUID SENSOR (液体センサ)] > [ENABLE (有効)] に進み、液体センサが有効になっていることを確認し ます。 2. 取入管をサンプル水または水道水に入れます。 注: 最高の精度を得るためにもサンプル水を使用してください。 3. ポンプの出口側から出ているチューブをメスシリンダーに入れます。 4. [VOLUME CALIBRATION (量の校正)] を押して、[CALIBRATION (校正)] を選択します。 5. [START (開始)] を選択します。ポンプが逆転して、取入管がパージされます。その後、ポンプが 正転して、サンプリングプログラムに指定されているサンプル量が採取されます。ポンプが逆転 して、取入管がパージされます。 6. サンプルの採取が完了したら [DONE (完了)] を選択します。メスシリンダー内のサンプル量と、 プログラム設定メニューに入力したサンプル量を比較します。 7. 採取された量が現在のプログラムに指定されているサンプル量と異なる場合は、実際に採取され た量を入力します。[ENTER (入力)] を押します。プログラムされているすべての量が正確に測 定されるようにセンサが調整されます。 8. [START (開始)] を選択してサンプリングプログラムを開始するか、[CANCEL (キャンセル)] を選 択して量の校正メニューを終了します。 サンプル量の手動校正 液体センサが無効な場合は、すべての量を時間によって手動校正する必要があります。サンプル量 は、現在のプログラムに指定されている量で校正されます。プログラムでサンプル量を変更した場合 は、新しいサンプル量で再度校正する必要があります。 1. [PROGRAM SETUP (プログラムの設定)] > [MODIFY SELECTED (選択修正)] > [LIQUID SENSOR (液体センサ)] > [DISABLE (無効)] を選択して、液体センサが無効になっていることを確 認します。 2. 取入管をサンプル水または水道水に入れます。 日本語 147 3. ポンプの出口側から出ているチューブをメスシリンダーに入れます。 4. [VOLUME CALIBRATION (量の校正)] を押します。複数の量が表示された場合は、校正するいず れかの量を選択します。 5. [START (開始)] を選択します。ポンプが逆転して、取入管がパージされます。その後、ポンプが 正転して、サンプルの採取が開始されます。 6. 選択した量または洗浄位置でポンプを停止します。 • サンプル量: メインプログラムに指定されている量が採取されたら [STOP (停止)] を押します。 • 洗浄: 取入管内の液体が液体検出器に達したら [STOP (停止)] を押します。 • 雨水量: 雨水プログラムに指定されている量が採取されたら [STOP (停止)] を押します。 7. [DONE (完了)] を選択して次の量の校正に進むか、[REPEAT (繰り返し)] を選択して再度校正を行 います。 8. すべての量の校正が完了したら、[DONE (完了)] を選択します。 9. 出口側のチューブをサンプラーのチューブ継手に接続します。 10. [START (開始)] を選択して、サンプリングプログラムを開始するか、[CANCEL (キャンセル)] を 選択して終了します。 サンプル量の確認 サンプル量が正確であるかどうかを確認するには、グラブサンプルを採取します。量を確認する目的 では校正に戻らないでください。そうすると、校正開始時に量の補正値がゼロにリセットされるから です。 1. 2. 3. 4. 5. 6. 7. [MANUAL OPERATION (手動操作)] を押します。 [GRAB SAMPLE (グラブサンプル)] を選択します。 取入管をサンプル水に入れます。 ポンプの出口側から出ているチューブをメスシリンダーに入れます。 液体センサが有効になっている場合は、確認する量を入力します。 液体センサが無効になっている場合は、プログラムに指定した量を入力します。 [ENTER (入力)] を押します。ポンプのサイクルが開始されます (パージ、サンプリング、パー ジ)。 8. メスシリンダー内の採取された量と想定していた量を比較します。採取された量が正しくない場 合は、量の校正を再度実行します。 メンテナンス 危険 複合的な危険。本書のこのセクションに記載されている作業は、必ず資格のある要員が行う必要があり ます。 危険 感電死の危険。メンテナンスまたはサービス活動を行う前に装置から電源を切り離します。 警告 生物学的危険への曝露。試料ボトルおよびサンプラーコンポーネントと接触している間は安全処理手 順に従ってください。 148 日本語 警告 複合的な危険。メンテナンス作業後には、機器が安全かつ正しく作動することを技術者が確認する必要 があります。 装置の清掃 注意 火災の危険。装置の洗浄に、可燃性の洗浄剤を使用しないでください。 告知 コントローラーコンパートメントヒーターは液体類を使って清掃しないでください。 水ではコントローラーとポンプを十分に清掃できない場合は、コントローラーの接続を外して、コン トローラーをサンプラーから取り外してください。コントローラーとポンプを完全に乾かしてから、 部品を取り付けて作動させてください。 サンプラーは次の手順で清掃します: • サンプラー筐体—湿った布と中性洗剤で内部および外部の表面を清掃してください。研磨剤入り 洗浄剤や溶剤は使用しないでください。 • サンプラーボトル—ブラシと水および中性洗剤でボトルを清掃してください。真水ですすいでか ら、蒸留水ですすいでください。 必要に応じて、加圧滅菌器でガラスボトルを清掃してください。 乾燥剤の交換 コントローラー内の乾燥剤カートリッジは、湿気を吸収し、腐食を防止します。点検窓を通して乾燥 剤の色を確認できます (図 12)。新品の乾燥剤の色は黄色です。緑色になったら乾燥剤を交換してく ださい。 1. 2. 3. 4. 5. 6. 乾燥剤カートリッジを回して取り外します (図 12)。 プラグを取り外して、使用済みの乾燥剤を廃棄します。 乾燥剤カートリッジに新品の乾燥剤を入れます。 プラグを取り付けます。 O リングにグリスを塗布します。 乾燥剤カートリッジをコントローラーに取り付けます。 図 12 乾燥剤カートリッジ 1 プラグ 3 O リング 2 乾燥剤カートリッジ 4 乾燥剤点検窓 日本語 149 ポンプのメンテナンス 注意 作業に際して、機器に指や皮膚を挟み込む可能性があります。メンテナンスまたは修理作業を完了する 前に、装置の電源を切ってください。 ポンプチューブの交換 告知 メーカーが供給したチューブ以外のチューブを使用すると、機械部品が過度に摩耗したり、ポンプの性能が低下 したりすることがあります。 ポンプチューブのローラーが当たる部分の摩耗を点検してください。通知メッセージが表示された 場合、またはチューブに摩耗の形跡が認められた場合は、チューブを交換してください 必要条件: • ポンプチューブ - プレカットまたはバルク 4.6 m または 15.2 m 1. コントローラーの電源を切り離します。 2. バルクチューブを使用する場合は、チューブをカットして、位置決め用のドットを付けます。 メ ーカーの Web サイトにある拡張版マニュアルを参照してください。 3. 図の番号に従ってポンプチューブを取り外します。 4. ポンプハウジングの内部とローラーに付着しているシリコン残留物を取り除きます。 5. 図の番号に従って新品のポンプチューブを取り付けます。 150 日本語 ローターの清掃 低刺激の洗剤を使用して、ローター、ポンプチューブ溝、およびポンプハウジングを清掃します。ポ ンプチューブの交換 ページの 150 とそれに続いて図示されている手順に従って取り付けてくださ い。 日本語 151 ディストリビューターアームチューブの交換 ディストリビューターアームは、複数ボトルのサンプリング時に各ボトルの上を移動します。ディス トリビューターアームのチューブが摩耗したら交換してください。ディストリビューターとディス トリビューターアームに適したチューブを使用してください。 注: ディストリビューターチューブはポンプチューブと異なります。ディストリビューターアセンブリにポンプチ ューブを取り付けると、ディストリビューターが損傷することがあります。また、ディストリビューターアームは 容易に動かせないので、試料が失われることもあります。 1. ディストリビューターアームと、センターセクションの天井からチューブを取り外します。 2. ディストリビューターアームに新品のチューブを取り付けます。チューブが 図 13 に示された寸 法に合っているか確認します。 3. チューブのもう一方の端をセンターセクションの天井にある継手に差し込みます。 4. ディストリビューターの診断テストを行って、正しく動作することを確認します。 図 13 ディストリビューターアセンブリ 1 ノズル 3 ディストリビューターモーター 2 ディストリビューターアーム 4 シャフト 5 ディストリビューターチューブ ヒューズの交換 ヒューズはユーザーが交換可能な部品ではありません。技術サポートにお問い合わせください。ヒ ューズが切れた場合は、サンプラーに問題が発生していて、点検整備が必要な可能性があります。 152 日本語 목차 사양 페이지의 153 작동 페이지의 166 일반 정보 페이지의 155 유지관리 페이지의 172 설치 페이지의 158 추가 정보 제조 회사의 웹 사이트에서 추가 정보를 확인할 수 있습니다. 사양 사양은 사전 예고 없이 변경될 수 있습니다. SD900 휴대용 샘플 채취기 사양 표준형 베이스 소형 베이스 복합형 베이스 규격 지름: 50.5 cm(19.9 in.) 지름: 44.1 cm(17.4 in.) 지름: 50.28 cm(19.8 in.) 높이: 69.4 cm(27.3 in.) 높이: 61 cm(24 in.) 높이: 79.75 cm(31.4 in.) 무게 15 kg(35.6 lb), 1 L 폴리에틸렌 병 24 개 12.2 kg(27 lb), 575 mL(19.44 oz) 폴리에틸렌 병 포함 24 개 포함 14.8 kg(32.6 lb), 10 L(2.5 gal) 폴리에 틸렌 용기 1 개 포함 15 kg(36 lb), 950 mL(32.12 oz) 폴리에틸렌 병 12 개 포함 12.9 kg(28.3 lb), 10 L(2.5 gal) 폴 리에틸렌 용기 1 개 포함 케이 스 3 중 구조의 충격 방지 ABS, 2.54 cm(1 in.)의 이중 벽구조 베이스 단열 처리 - 얼음에 직접 접촉 가능한 병 샘플 온도 0~60 °C(32~140 °F) 여과 기 표준 규격의 316 스테인리스 스틸, 저심도 채취용 고속 또는 저속 프로필, Teflon® 또는 표준 규격의 316 스 테인리스 스틸 샘플 흡입 배관 9.5 mm(3/8 in.) I.D 비닐 또는 Teflon 라이닝 폴리에틸렌 샘플 1 L(0.26 gal) 폴리에틸렌 병 및/또는 병 용 350 mL(11.83 oz) 유리병 24 개 량 2.3 L(0.6 gal) 폴리에틸렌 병 및/또는 1.9 L(0.5 gal) 유리병 8 개 3.8 L(1 gal) 폴리에틸렌 병 및/또는 3.8 L(1 gal) 유리병 4 개 575 mL(19.44 oz) 폴리에틸렌 병 24 개 21 L(5.5 gal) 폴리에틸렌 병 1 개 950 mL(32.12 oz) 유리병 8 개 10 L(2.5 gal) 폴리에틸렌 병 1 개 10 L(2.5 gal) 유리병 1 개 3.8 L(1 gal) 폴리에틸렌 병 및/또는 3.8 L(1 gal) 유리병 2 개 21 L(5.5 gal) 또는 15 L(4 gal) 폴리에틸 렌 합성 용기 또는 20 L(5.25 gal) 폴리 에틸렌 병 또는 10 L(2.5 gal) 폴리에틸 렌 병 또는 10 L(2.5 gal) 유리병 1 개 SD 컨트롤러 사양 세부 사항 치수 (W x H x D) 29.2 x 17.1 x 26.4 cm (11½ x 6¾ x 103/8 in.) 무게 4.2 kg (9.26 lb) 한글 153 사양 세부 사항 케이스 PC/ABS 혼합, NEMA 4X, 6, IP 67 전원 요구 사항 8754500 전원공급장치로 공급되는 15 VDC, 내부 전원공급장치로 공급되는 15 VDC 과부하 보호 펌프용 7 A, DC 라인 퓨즈 작동 온도 0-50°C(32-122°F) 보관 온도 –30-60 °C (–22-140 °F) 보관/작동 습도 100% 응결 펌프 튜브 연동 고속, 스프링 장착식 Nylatron 롤러 펌프 케이스 IP37 펌프 배관 9.5 mm ID x 15.9 OD mm (3/8 in. ID x 5/8 in. OD) 실리콘 펌프 배관 수명 다음 조건에서 20,000 샘플 사이클: 1 L (0.3 gal) 샘플 볼륨, 1 회 린스, 6 분 페이싱 주기, 4.9 m (16 ft) 길이 3/8 in. 흡입 튜브, 4.6 m (15 ft) 수직 양정, 21 °C (70 °F) 샘플 온도 샘플 수직 양정 다음 조건에서 최대 8.5 m (28 ft): 8.8 m (29 ft) 길이 3/8-in. 비닐 흡입 튜브, 해발, 20–25 °C (68–77 °F) 펌프 유량 4.8 L/min (1.25 gpm), 3/8-in. 흡입 튜브를 사용하고 1 m (3 ft) 수직 양정 시 (일반) 시료 볼륨 10-10,000 mL (3.38 oz-2.6 gal) 범위에서 10 mL (0.34 oz) 단위로 프로그래밍 가능 시료 볼륨 반복성 (일 반) 다음 조건에서 무교정 액체 검출 시 200 mL 샘플 볼륨의 ±5%: 4.6 m (15 ft) 수직 양정, 4.9 m (16 ft) 길이 3/8-in. 비닐 흡입 튜브, 단일 용기, 용기 찰 경우 중단, 실온 및 1524 m (5000 ft) 고도 시료 볼륨 정확도 (일 반) 다음 조건에서 무교정 액체 검출 시 200 mL 샘플 볼륨의 ±10%: 4.6 m (15 ft) 수직 양정, 4.9 m (16 ft) 길이 3/8-in. 비닐 흡입 튜브, 단일 용기, 용기 찰 경우 중단, 실온 및 1524 m (5000 ft) 고도 샘플링 모드 속도: 시간 고정, 유량 고정, 시간 가변, 유량 가변, CTVV(일정 시간 동안 변화 가능 볼륨) CTVV(일정 시간 동안 변화 가능 볼륨)은 본 설명서의 확장 버전을 참조하십시오. 분산: 단일 병에 합성, 여러 병에 합성, 여러 병에 분리, 샘플 당 병, 병 당 샘플, 샘플 당 병 의 조합, 병 당 샘플 실행 모드 연속식 또는 비연속식 (사용자가 샘플 수 입력) 다중 프로그램 최대 3 개의 샘플 채취 프로그램 저장 이송 속도 (일반) 다음 조건에서 0.9 m/s (2.9 ft/s): 4.6 m (15 ft) 수직 양정, 4.9 m (16 ft) 길이 3/8-in. 비닐 흡 입 튜브, 21 °C (70 °F) 및 1524 m (5000 ft) 고도 액정 센서 초음파. 본체: Ultem NSF ANSI 표준 51 승인, USP Class VI 준수 샘플 기록 최대 510 개 기록 공기 정화 각 샘플 전후에 자동으로 공기 정화. 지속 시간이 자동으로 흡입 튜브 길이 변화를 보정. 이벤트 로그 510 개 기록 연결 전원, 보조, 시리얼 통신, 디스트리뷰터, SDI-12 습식 재질 스테인리스 스틸, 폴리에틸렌, Teflon, Ultem, 실리콘 통신 RS232, Modbus, SDI-12 보증 1년 154 한글 ® 일반 정보 제조업체는 본 설명서에 존재하는 오류나 누락에 의해 발생하는 직접, 간접, 특수, 우발적 또는 결과적 손해에 대해 어떠한 경우에도 책임을 지지 않습니다. 제조업체는 본 설명서와 여기에 설명된 제품을 언제라도 통지나 추가적 책임 없이 변경할 수 있습니다. 개정본은 제조업체 웹 사이트에서 확인할 수 있습니다. 안전 정보 주의사항 제조사는 본 제품의 잘못된 적용 또는 잘못된 사용으로 인한 직접, 우발적 또는 간접적 손해에 국한하지 않는 모든 손해에 대한 어떠한 책임도 지지않으며, 관계 법령이 최대한 허용하는 손해에 관한 면책이 있습니다. 사용자는 사 용상 중대한 위험을 인지하고 장비 오작동이 발생할 경우에 대비하여 적절한 보호 장치를 설치하여야 합니다. 장치 포장을 풀거나 설치하거나 작동하기 전에 본 설명서를 모두 읽으십시오. 모든 위험 및 주의사항 설명에 유의하시기 바랍니다. 이를 지키지 않으면 사용자가 중상을 입거나 장치가 손상될 수 있습니 다. 본 장치의 보호 기능이 손상되지 않도록 본 설명서에서 설명하는 방법이 아닌 다른 방법으로 본 장치 를 사용하거나 설치하지 마십시오. 위험 정보 표시 위험 방지하지 않을 경우 사망 또는 심각한 부상이 일어나는 잠재적 또는 즉각적 위험 상황을 의미합니다. 경고 피하지 않을 경우에 사망이나 심각한 부상을 유발할 수 있는 잠재적 위험이나 긴급한 위험 상황을 나타냅니다. 주의 경미하거나 심하지 않은 부상을 초래할 수 있는 잠재적으로 위험한 상황을 경고합니다. 주의사항 피하지 않으면 기기에 손상을 일으킬 수 있는 상황을 나타냅니다. 특별히 강조할 필요가 있는 정보. 주의 경고 본 기기에 부착된 모든 라벨 및 태그를 참조하시기 바랍니다. 지침을 따르지 않을 경우 부상 또는 기기 손상이 발생할 수 있습니다. 기기에 있는 기호는 주의사항에 대한 설명과 함께 설명서에서 참조합니 다. 이는 안전 경고 심볼입니다. 잠재적인 부상 위험을 방지할 수 있도록 이 기호를 따라 모든 안전 메시지 를 준수하십시오. 기기에 안전 기호가 부착되어 있는 경우 작동 및 안전 정보에 대해서는 작동 설명서를 참조하십시오. 본 심볼은 잠재적인 압착 위험을 나타냅니다. 이 심볼이 표시된 전기 장비는 유럽 내 공공 폐기 시스템에 따라 폐기할 수 없습니다. 인증 캐나다 무선 간섭 유발 장치 규정, IECS-003, 등급 A: 보조 테스트 기록은 제조업체가 제공합니다. 한글 155 본 등급 A 디지털 장치는 캐나다 간섭 유발 장치 규제의 모든 요구조건을 만족합니다. Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences. FCC Part 15, Class "A" 제한 보조 테스트 기록은 제조업체가 제공합니다. 본 장치는 FCC 규칙, Part 15 를 준수합니다. 본 장치는 다음 조건에 따라 작동해야 합니다. 1. 유해한 간섭을 일으키지 않아야 합니다. 2. 바람직하지 않은 작동을 유발할 수 있는 간섭을 포함하여 수신되는 모든 간섭에도 정상적으로 작 동해야 합니다. 본 장치의 준수 책임이 있는 측이 명시적으로 허용하지 않은 변경 또는 수정을 가하는 경우 해당 사용 자의 장치 작동 권한이 무효화될 수 있습니다. 본 장치는 FCC 규칙, Part 15 에 의거하여 등급 A 디지 털 장치 제한 규정을 준수합니다. 이러한 제한은 상업 지역에서 장치를 작동할 때 유해한 간섭으로부 터 적절하게 보호하기 위하여 제정되었습니다. 본 장치는 무선 주파수 에너지를 생성 및 사용하며 방 출할 수 있고 사용 설명서에 따라 설치하고 사용하지 않을 경우 무선 통신에 해로운 간섭을 일으킬 수 있습니다. 주거 지역에서 본 장치를 사용하면 해로운 간섭을 일으킬 수 있으며, 이 경우 사용자는 자비 를 들여 간섭 문제를 해결해야 합니다. 다음과 같은 방법으로 간섭 문제를 줄일 수 있습니다. 1. 2. 3. 4. 5. 장치를 전원에서 분리하여 장치가 간섭의 원인인지 여부를 확인합니다. 장치가 간섭을 받는 장치와 동일한 콘센트에 연결된 경우, 장치를 다른 콘센트에 연결해보십시오. 장치를 간섭을 받는 장치로부터 멀리 떨어트려 놓으십시오. 간섭을 받는 장치의 안테나 위치를 바꿔보십시오. 위의 방법들을 함께 이용해보십시오. 제품 개요 위험 화학적 또는 생물학적 위험 존재. 본 장비를 공중 위생, 공중 안전, 식음료 제조 또는 가공에 관련한 시행 령 및 감시 규정 목적으로 처리공정이나 약품 주입 시스템을 감시하기 위하여 사용하는 경우, 이 장비에 적용되는 모든 규정을 이해하고 준수하며, 장비가 오작동하는 경우 해당 규정에 따라 충분하고 합당한 메커니즘을 보유하는 것은 사용자의 책임입니다. 주의 화재 위험. 본 제품은 가연성 액체를 사용할 수 있게 설계되지 않았습니다. SD900 휴대용 샘플 채취기는 지정된 간격으로 액체 샘플을 채취하여 병 또는 용기에 보관합니다. 샘 플 채취기는 다양한 수성 샘플을 비롯하여 독성 오염 물질 및 부유 물질 채취 용도로 사용할 수 있습니 다. 샘플 채취기는 여러 리테이너, 병 또는 용기와 함께 설치할 수 있습니다. 그림 1 을 참조하십시오. 156 한글 그림 1 제품 개요 1 소형 베이스 6 컨트롤러 연결 2 표준형 단열 베이스 7 펌프 3 중심부 8 컨트롤러 4 전원 9 액정 센서 5 상단 덮개 10 21 L(5.5 gal) 컨테이너용 표준형 단열 베이스 제품 구성 모든 구성품을 수령했는지 확인하십시오. 그림 2 를 참조하십시오. 품목이 누락되었거나 손상된 경우 에는 제조업체 또는 판매 담당자에게 즉시 연락하십시오. 그림 2 제품 구성 1 베이스(표준형, 소형 또는 복합형) 6 배터리 충전기(옵션) 2 단일 병으로 채취 시의 부속품(병 및 지지대는 변경 될 수 있음) 7 AC 전원공급장치(옵션) 3 중심부(컨트롤러) 8 여과기 4 배터리(옵션) 9 흡입 튜브, 비닐 또는 Teflon 라이닝 5 상단 덮개 10 여러 병으로 채취 시의 부속품(병 및 리테이너는 변 경될 수 있음) 한글 157 설치 위험 여러 가지 위험이 존재합니다. 해당 전문가가 본 문서에 의거하여 작업을 수행해야 합니다. 밀폐 공간 주의사항 위험 폭발 위험 XXX 다음 정보는 사용자가 밀폐 공간에 들어갈 때 발생할 수 있는 위험을 이해하도록 돕기 위해 제공되는 것입니다. 1993 년 4 월 15 일, OSHA 의 CFR 1910.146, 작업 허가 필요 밀폐 공간에 대한 최종 판결이 입법되었 습니다. 이 표준은 미국 내 250,000 여 사이트에 직접적인 영향을 미치며, 밀폐 공간에서의 작업자 보 건 및 안전을 보호하기 위해 제정되었습니다. 밀폐 공간의 정의: 밀폐 공간은 다음 중 하나 이상의 조건을 가지는 (또는 그럴 가능성이 임박한) 위치 또는 외함입니다. • 산소 농도가 19.5% 미만 또는 23.5% 초과 및/또는 황화수소(H2S) 농도가 10 ppm 초과하는 대기 • 가스, 증기, 연무, 분진 또는 파이버로 인해 가연성 또는 폭발성이 될 수 있는 대기 • 접촉 또는 흡입 시 부상, 건강 유해성 또는 사망을 초래할 수 있는 독성 물질 밀폐 공간은 인간이 점유하도록 설계된 것이 아닙니다. 밀폐 공간은 출입이 제한되며, 알려진 또는 잠 재적인 위험을 가지고 있습니다. 밀폐 공간의 예로는 맨홀, 스택, 파이프, 대형 통, 스위치 볼트 및 기타 유사한 위치가 포함됩니다. 밀폐 공간 및/또는 유해 가스, 증기, 연무, 분진 또는 파이버가 존재할 수 있는 위치에 들어가기 전에 항 상 표준 안전 절차를 준수해야 합니다. 밀폐 공간에 들어가기 전에 밀폐 공간 출입과 관련된 모든 절차 를 찾아 숙지하십시오. 기계 설치 사이트 설치 지침 위험 폭발 위험. 본 기기는 위험 위치에 설치하도록 승인되지 않았습니다. 설치 장소 평가에 대해서는 지침을 참조하십시오. • 샘플 채취기를 좁은 공간에 설치한 경우 모든 안전 수칙을 따르십시오. 밀폐 공간 주의사항 페이지의 158 을 참조하십시오. • 설치 장소의 기온은 명시된 범위를 유지해야 합니다. 사양 페이지의 153 을 참조하십시오. • 샘플 채취기를 평평한 지면에 설치하거나 서스펜션 하니스, 서포트 브래킷 또는 스패너 바로 샘플 채취기를 지탱해야 합니다. 맨홀 안에 샘플 채취기 설치 페이지의 159 및 적절한 설치 지침 문서를 참조하십시오. • 측정 지연 시간을 줄이기 위해 최대한 샘플 소스와 가까워야 합니다. 샘플 채취기 배관 연결 페이지의 161 을 참조하십시오. • 운송 속도 제한 및 최대 수직 양정 제한에 대한 정보는 사양 페이지의 153 을 참조하십시오. 158 한글 맨홀 안에 샘플 채취기 설치 맨홀 안에 있는 시료수 위에 샘플 채취기를 설치합니다. 스패너 바 또는 서포트 브래킷으로 샘플 채취 기를 설치합니다. 맨홀 내부에 스패너 바를 설치합니다. 스패너 바는 양쪽 벽을 미는 힘으로 지지합니 다. 서포트 브래킷의 폭은 맨홀 뚜껑과 같습니다. 서포트 브래킷은 뚜껑 바로 밑에 설치하여 지지합니 다.제조업체 웹 사이트의 세부 설명서를 참조하십시오.샘플 채취기샘플 채취기. 디스트리뷰터 또는 풀-보틀(full-bottle) 차단장치(옵션) 디스트리뷰터 또는 풀-보틀 차단장치는 일반적으로 공장에서 설치됩니다. 설치 시 디스트리뷰터 또는 풀-보틀 차단 장치 문서를 참조하십시오. 샘플 채취기 준비 샘플 병 청소 브러시, 물, 연성 세제로 샘플 병과 마개를 청소합니다. 병을 깨끗한 물로 씻은 다음 증류수로 헹굽니 다. 단일 용기 설치 1 개의 병을 사용하여 하나의 복합 샘플을 채취합니다. 병이 가득 차면 용기 찰 경우 중단 기능이 샘플 프로그램을 중지합니다. 1. 샘플 병을 청소합니다. 샘플 병 청소 페이지의 159 을 참조하십시오. 2. 그림 3 에 나온 대로 샘플 병을 설치합니다. 그림 3 단일 용기 설치 1 10 L(2.5 gal) 폴리에틸렌 병 (1918) 6 21 L(5.5 gal) 폴리에틸렌 병 (6494) 2 10 L(2.5 gal) 유리병 (6559) 7 소형 베이스 (8975) 3 지지대 (1502) 8 표준형 단열 베이스 (8976) 4 15 L(4 gal) 폴리에틸렌 병 (1367) 9 21 L(5.5 gal)용 표준형 단열 베이스 (8561) 5 19 L(5 gal) 폴리에틸렌 병 (6498) 여러 샘플 채취 시 첫 번째 병 설치 여러 개의 병을 사용하여 두 개 이상의 병에 샘플을 각각 채취합니다. 디스트리뷰터는 각 병 위에서 샘 플 튜브를 움직입니다. 그림 4 에 나와 있는 것과 같이 병을 샘플 채취기 베이스에 설치합니다. 첫 번째 샘플 병(1 번)을 샘플 채취기 베이스의 라벨 아래에 설치합니다. 남아 있는 병을 라벨에 나와 있는 숫자 의 순서대로 설치합니다. 필수 구성 요소의 도표는 그림 5 페이지의 161 를 참조하십시오. 한글 159 그림 4 1 번 병 설치 1 1 번 병 위치(소형 베이스) 6 고무줄 2 병 24 개 사용 시 1 번 병 위치 7 21 L(5.5 gal)용 표준형 단열 베이스 3 병 8 개 사용 시 1 번 병 위치 8 표준형 단열 베이스 4 병 2 개 또는 4 개 사용 시 1 번 병 위치 9 소형 베이스 5 리테이너 다중 용기 설치 여러 개의 병을 설치할 경우 디스트리뷰터 암이 샘플 튜브를 각 병 위로 이동시킵니다. 지정된 수의 샘 플이 채취되면 샘플 채취가 자동으로 중지합니다. 1. 샘플 병을 청소합니다. 샘플 병 청소 페이지의 159 을 참조하십시오. 2. 첫 번째 샘플 병(1 번)을 샘플 채취기 베이스의 라벨 아래에 설치합니다. 여러 샘플 채취 시 첫 번째 병 설치 페이지의 159 를 참조하십시오. 3. 그림 5 에 나온 대로 샘플 병을 조립합니다. 병이 8 개 이상일 경우 첫 번째 병이 시계 방향으로 1 번 병 표시기 옆에 오는지 확인합니다. 4. 병 어셈블리를 시료 채취기에 놓습니다. 병이 8 개 이상일 경우 와이어를 하단 트레이의 슬롯과 맞 춥니다. 160 한글 그림 5 다중 용기 설치 1 포지셔너/리테이너 (2347) 8 3.8 L(1 gal) 유리병 4 개 (2216) 또는 3.8 L(1 gal) 폴 리에틸렌 병 4 개 (2217) 2 리테이너 (1422) 9 1.9 L(0.5 gal) 유리병 8 개 (1118) 또는 2.3 L(0.61 gal) 폴리에틸렌 병 8 개 (657) 3 리테이너 (2190) 10 1 L(0.26 gal) 폴리에틸렌 병 24 개 (737) 4 리테이너 (2189) 11 350 mL(11.83 oz) 유리병 24 개 (732) 5 950 mL(32.12 oz) 유리병 8 개 (2348) 12 소형 베이스 (8975) 6 575 mL(19.44 oz) 폴리에틸렌 병 24 개 (1369) 13 표준형 베이스 (8976) 7 3.8 L(1 gal) 유리병 2 개 (2214) 또는 3.8 L(1 gal) 폴 리에틸렌 병 2 개 (2215) 샘플 채취기 배관 연결 대표 샘플이 채취되도록 (표면 또는 바닥 가까이가 아니라) 시료 흐름의 중앙에 여과기를 설치합니다. 그림 6 을 참조하십시오. 흡입 튜브 설치 시 그림 7 을 참조하십시오. 1. 그림 7 에 나온 대로 시료 채취기에 튜브를 연결합니다. 참고: Teflon 라이닝 튜브를 사용하는 경우 연결 키트 2186 을 사용합니다. 2. 흡입 튜브 및 여과기를 물이 요동치고 잘 섞이는 샘플 소스의 본류에 설치합니다. • 흡입 튜브는 가능한 한 짧게 만듭니다. • 샘플 사이에 흡입 튜브가 완전히 배수되도록 튜브를 최대한 수직으로 유지합니다. 참고: 수직이 불가능하거나 튜브가 가압되는 경우 액정 센서를 사용하지 마십시오. 시료 볼륨을 수동으로 교 정하십시오. 시료 볼륨을 수동으로 교정 페이지의 171 을 참조하십시오. • 흡입 투브가 집히지 않았는지 확인합니다. 한글 161 그림 6 기기 설치 1 여과기 3 흡입 튜브 2 수직 양정 그림 7 흡입 튜브 설치 전기 설치 컨트롤러 연결 경고 전기쇼크 위험. 외부 연결된 장비는 해당하는 국가 안전 표준에 따라 평가를 마쳐야 합니다. 그림 8 에는 컨트롤러에 연결할 수 있는 커넥터가 나와 있습니다. 162 한글 그림 8 컨트롤러 커넥터 1 보조 장치 4 SDI-12 장치 옵션 2 전원공급장치 5 디스트리뷰터/용기 찰 경우 중단 3 시리얼 통신 유량계 연결 유량계를 컨트롤러에 연결하여 샘플 유량이 지정된 값을 상회 또는 하위하면 샘플 채취기를 시작하거 나 중지합니다. 준비 항목: • Sigma 유량계용 다목적 풀 케이블 (또는 모델 980 유량계용 980 하프 케이블) • 추가 연결용 분배기 (옵션). 2 개 이상의 분배기를 직렬로 연결할 수 있습니다. 1. 케이블의 한쪽 끝을 유량계에 연결합니다. 모델 980 유량계의 경우 모델 980 유량계 사용 설명서 를 참조하십시오. 2. 케이블의 다른 쪽 끝을 컨트롤러의 보조 장치 커넥터에 연결합니다. 참고: 유량계가 6 핀 케이블을 사용할 경우 6 핀-7 핀 어댑터 케이블을 사용하십시오. PC 또는 통신 네트워크 연결 데이터를 전송하거나 샘플 채취기를 구성하려면 PC 또는 통신 네트워크를 컨트롤러에 연결합니다. 준비 항목: • 시리얼 케이블, 7 핀 RS232 - DB-9 (그림 9 참조). 연결: B - 5 (신호 접지), D - 3 (RCD), F - 2 (TXD), G (접지) 1. 시리얼 케이블의 한쪽 끝을 통신 장치 또는 네트워크에 연결합니다. 2. 케이블의 다른 쪽 끝을 컨트롤러의 시리얼 통신 커넥터에 연결합니다. 그림 9 시리얼 통신 케이블 한글 163 샘플 채취기 전원 연결 위험 전기 쇼크 위험. 기기를 실외 또는 습기있는 장소에서 사용하는 경우, 접지 결함 회로 인터럽트 (GFCI/GFI) 장치를 사용하여 기기를 메인 전원에 연결 합니다. 위험 화재 위험. 전원 라인에 15 A 용량의 회로 차단기를 설치하십시오. 회로 차단기는 장비 근처에 설치했을 때 국부 전원 차단 역할을 합니다. 위험 전기쇼크 위험 보호 접지(PE) 연결이 필요합니다. 경고 전기쇼크 위험 해당 위치에서 전원을 쉽게 차단할 수 있어야 합니다. 경고 감전 위험. 펌프 사이클 사이의 시간이 너무 짧으면 전원 공급기가 과열될 수 있습니다. 주의사항 펌프 사이클 시간 때문에 전원공급장치가 과열되지 않아야 합니다. 펌프 사이클 시간 계산 페이지의 165 을 참조하 십시오. 샘플 채취기에 배터리 또는 AC 전원공급장치를 연결합니다. 아래의 단계별 그림 설명을 참조하십시 오. 164 한글 펌프 사이클 시간 계산 전원공급장치가 과열되지 않도록 샘플 사이클 사이에 펌프가 최소 기간 동안 작동을 중지해야 합니다. 1. 샘플 사이클 동안 펌프가 연속 작동할 총 시간을 계산합니다. 다음 단계를 모두 포함합니다. 정화 전, 흡입 튜브 린스, 샘플, 샘플 재시도 및 정화 후. 2. 그림 10 에서 샘플 사이클 사이에 펌프가 작동을 중지해야 하는 최소 시간을 찾습니다. 3. 샘플 채취기 프로그램의 페이싱 또는 시간 주기가 샘플 사이클 사이에 최소 시간 동안 펌프를 작동 중지 상태로 유지하는지 확인합니다. 그림 10 50 °C (122 ºF) 시 펌프 작동 및 중지 횟수 1 펌프가 작동 중지해야 하는 시간(분) 2 펌프가 연속 작동하는 시간(분) 3 예: 펌프가 5 분간 연속 작동할 경우 펌프는 15 분간 작동을 중지해야 합니다. 전원 켜기 전원 버튼을 눌러 전원을 켜거나 끄십시오. 전원공급장치(AC 전원 또는 배터리)가 올바르게 설치됐는 지 확인하십시오. 사용자 인터페이스 및 탐색 컨트롤러 키패드는 그림 11 에 나와 있습니다. 샘플 채취 프로그램을 설정하거나 컨트롤러 설정을 구 성하거나 진단 시험을 실시하려면 MENU(메뉴) 키를 사용합니다. 메뉴를 스크롤하고, 항목을 선택하 고, 값을 입력하려면 화살표, ENTER(엔터) 및 BACK(뒤로) 키를 사용합니다. 더 많은 옵션을 사용할 수 있는 경우 디스플레이 화면의 하단 또는 측면에 화살표 키가 표시됩니다. 한글 165 그림 11 컨트롤러 키패드 1 POWER(전원) 5 메뉴 9 상태 2 VOLUME CALIBRATION(볼륨 교정) 6 ENTER 10 RUN/HALT PROGRAM(프로그 램 실행/중단) 3 정지 7 뒤로 4 LED 8 화살표 키 11 MANUAL OPERATION(수동 작 동) 작동 시스템 설정 구성 1. MENU(메뉴)를 누릅니다. 2. SYSTEM SETUP(시스템 설정)으로 이동합니다. 3. 해당 옵션을 변경합니다. 옵션 설명 시간/날짜 설정 시간 (24 시간 형식) 및 날짜를 설정합니다. 통신 시리얼 포트의 전송 속도 (19200, 38400, 57600 또는 115200) 및 프로토콜 (Modbus RTU 또는 ASCII)을 선택합니다. 기본 설정 샘플 채취기 베이스를 선택합니다(전천후 냉장형, 냉장형, 휴대용 표준형 또는 휴대용 소 형). 언어 컨트롤러 언어를 선택합니다. 액정 센서 교정 액정 센서를 교정하거나 기본값으로 설정합니다. 명암비 설정 LCD 화면의 명암비를 조정합니다. 펌프 듀티 사이클 펌프 듀티 사이클을 변경합니다. 범위: 50~100%(기본값: 100%) 암호 설정 166 한글 프로그램 설정 및 시스템 설정 메뉴에 대한 액세스를 제한하는 암호를 설정합니다. ENABLE(사용)>YES(예)를 선택하여 새 암호를 설정하거나 ENABLE(사용)>NO(아니요)를 선택하여 현재 암호를 활성화합니다. 암호를 최초로 설정하는 경우 현재 암호로 900900 을 사용하십시오. 옵션 설명 튜브 수명 펌프 튜브 교체 시기를 안내하는 알림 메시지를 설정합니다. ENABLE(사용)>RESET CYCLES(사이클 재설정)으로 이동하여 펌프 사이클 카운트를 0 으로 재설정합니다. ENABLE(사용)>CYCLE LIMIT(사이클 한계)로 이동하여 알림 메시지에 사용할 펌프 사이클 수를 변경합니다. ENABLE(사용)>STATUS(상태)로 이동하여 펌프 사이클 카운트의 현재 수치를 표시합니다. SDI–12 (옵션) SDI–12 장치의 작동을 구성합니다. 이 옵션은 장치가 발견될 때에만 표시됩니다. 샘플 채취 프로그램 설정 샘플 채취 프로그램에는 자동으로 샘플을 채취하는 데 필요한 모든 매개변수가 포함됩니다. 정기적으 로 또는 샘플 채취 프로그램이 완료되면 샘플을 채취합니다. 1. MENU(메뉴)를 누릅니다. 2. PROGRAM SETUP(프로그램 설정)>MODIFY ALL(모두 수정)으로 이동합니다. 첫 번째 화면이 표 시됩니다. 3. 각 매개변수의 값을 선택하거나 입력합니다. 왼쪽 또는 오른쪽으로 이동하려면 왼쪽/오른쪽 화살 표 키를 사용합니다. 값을 변경하려면 위쪽/아래쪽 화살표 키를 사용합니다. 필요할 경우 화살표 키 를 사용하여 단위 (gal 또는 mL)를 변경합니다. 옵션 용기 설명 용기 수: 샘플 채취기에 설치된 병의 개수(1, 2, 4, 8 또는 24). 용기 찰 경우 중단: 사용/사용 안 함. 사용으로 설정할 경우 컨트롤러가 병이 넘치는지 여부를 점검 합니다. 병 용량: 각 병의 용량. 범위: 50~65000 mL(0.01~17.17 gal) 흡입 튜 브 튜브 길이: 여과기에서 액정 센서까지의 흡입 튜브 길이. 범위: 100–3000 cm 또는 3–99 ft. 정확한 샘플 볼륨을 얻으려면 정확한 길이가 필요합니다. 흡입 튜브 유형: 흡입 튜브의 직경 및 재질 (¼ in. 비닐, 3/8 in. 비닐 또는 3/8 in. Teflon). 프로그램 사용/사용 안 함: 사용으로 설정할 경우 지정된 시간 또는 카운트 수에서 샘플 채취 프로그램이 시작 지연 합니다. 날짜 및 시간: 프로그램이 시작하는 날짜 및 시간 (24 시간 형식). 카운트: 프로그램이 시작 할 유량계 카운트 또는 펄스 수 (1-9999 카운트). 나중에 샘플 페이싱을 시간으로 설정할 경우 프로 그램 지연이 비활성화됩니다. 샘플 페 이싱/수 집 시간 기반 - 시간 기반 수집 페이지의 168 을 참조하십시오. 참고: CTVV 가 사용되는 경우: • • • • • 액정 센서가 항상 활성화됩니다. 샘플 분산이 합성됩니다. 실행 모드가 계속되지 않습니다. 첫 번째 샘플이 간격 이후 수집됩니다. 유량 소스 입력은 항상 4–20 mA 입니다. 유량 기반 - 유량 기반 수집 페이지의 169 을 참조하십시오. 샘플 분 배 샘플을 모든 병으로 분배>예 각 샘플이 모든 병으로 분배됩니다. 마지막 샘플 후 종료: 프로그램이 마지막 샘플 후 종료합니다 (1–999). 연속: 프로그램이 수동으로 중지될 때까지 계속됩니다. 샘플을 모든 병으로 분배>아니요 샘플이 지정된 그룹의 병으로 분배됩니다. . 샘플/용기: 각 병에 채취될 샘플 수 (1–999). 용기/샘플: 동일한 샘플이 채취될 병의 개수. 액정 센 서 사용 또는 사용 안 함. 사용 안 함으로 설정할 경우 샘플 볼륨을 시간 기반으로 교정해야 합니다. 시료 볼 륨 각 샘플의 볼륨 (단위: mL (1 gal = 3785.4 mL)). 용기/샘플 모드를 사용할 경우 각 병에는 전체 샘플 볼륨이 들어갑니다 (10–10,000 mL). 샘플 볼륨이 병 용량을 초과하지 않는지 확인하십시오. 샘플 볼륨은 10 mL 단위로 반올림합니다. 한글 167 옵션 설명 흡입 튜 브 린스 샘플 채취 전 흡입 튜브 린스 횟수 (0–3). 샘플 재 시도 횟 수 샘플 채취가 실패할 경우 샘플 채취 시도 횟수 (0–3). 사이트 ID 샘플 채취 위치의 이름 (최대 12 자). 프로그램이 저장되면 사이트 ID 가 사전 설정 이름으로 사용됩 니다. 고급 샘 플 채취 고급 샘플 채취 메뉴로 이동합니다. 필요할 경우 고급 샘플 채취 메뉴에서 옵션을 구성합니다. 프로그램 기본 샘플 프로그램 생성이 완료됩니다. 샘플 채취기가 프로그램을 시작 또는 취소하라는 메시지를 실행 표시합니다. 시간 기반 수집 1. 옵션의 값을 선택하거나 입력합니다. 옵션 설명 속도 간격 샘플을 일정한 시간 간격(단위: 시간 및 분 단위)으로 수집합니다 (0:01–999:00). 첫 번째 샘플 수 집 프로그램은 즉시 시작하거나 첫 번째 간격 이후에 시작합니다. 변화 가능한 볼 륨 샘플 볼륨의 변화 여부를 지정합니다(유량 기반). 변화할 경우 변화 가능한 볼륨 사용 페이지의 168 을 참조합니다. 변화하지 않을 경우 변화 가능한 볼륨을 비사용 페이지의 169 을 참조합니다. 변화 가능한 볼륨 사용 샘플 속도는 일정 시간 동안 변화 가능 볼륨(CTVV) 입니다. 1. 옵션의 값을 선택하거나 입력합니다. 옵션 설명 단위 선택 사용할 유량 단위를 선택합니다. MAP 4 mA 입력 4 mA 입력에 일치하는 유속을 입력합니다. MAP 20 mA 입력 20 mA 입력에 일치하는 유속을 입력합니다. 평균 유속 사이트의 평균 유속을 지정합니다(1–999,999). 단위는 위에서 선택한 단위로 지정됩니다. 원하는 시간 샘플 프로그램 도중 수집하고자 하는 총 볼륨을 지정합니다(10–10,000 ml). 168 한글 옵션 설명 수집 기간 수집 기간을 시간과 분으로 지정합니다(0:01–999:00). 최소 샘플량 유량이 최소 볼륨을 수집하기에 부족할 경우 샘플이 생략됩니다(10–10,000 ml). 변화 가능한 볼륨을 비사용 1. 옵션의 값을 선택하거나 입력합니다. 옵션 설명 첫 번째 샘플 수집 프로그램을 즉시 시작할 지, 첫 번째 간격 이후에 시작할 지 선택합니다. 유량 기반 수집 유량 기반 샘플의 흐름 소스를 다음과 같이 지정합니다. • 카운트 - 유량 소스에 카운트 사용 페이지의 169 을 참조하십시오. • 4–20 mA 입력 - 유량 소스에 4–20 mA 입력 페이지의 169 을 참조하십시오. 유량 소스에 카운트 사용 1. 옵션의 값을 선택하거나 입력합니다. 옵션 설명 매 간격마다 샘플 수집 일정 간격(단위: 카운트)으로 샘플을 채취합니다 (1–9999 카운트). 중지 시간 유량이 너무 적을 경우 샘플을 채취합니다. 시간 샘플 간 최대 시간 (0:01–999:00). 각 샘플이 수집되면 타이머가 다시 시작합니다. 첫 번째 샘플 수집 프로그램이 즉시 시작하거나 첫 번째 간격 이후에 시작합니다. 유량 소스에 4–20 mA 입력 1. 옵션의 값을 선택하거나 입력합니다. 옵션 설명 단위 선택 사용할 유량 단위를 선택합니다. MAP 4 mA 입력 4 mA 에 일치하는 유속을 입력합니다. MAP 20 mA 입력 20 mA 에 일치하는 유속을 입력합니다. 매 유량마다 샘플 수 집 샘플을 수집할 총 유량을 입력합니다. 중단 시간 유량 볼륨이 비정상적으로 낮을 때 샘플을 수집하려면 선택합니다. 시간 샘플 간 최대 시간 (0:01–999:00). 각 샘플이 수집되면 타이머가 다시 시작합니다. 시료 볼륨 샘플 당 수집할 볼륨을 입력합니다. 병/샘플 모드를 사용할 경우 각 병에는 전체 샘플 볼륨이 들어갑니다 (100–10,000 mL). 변화 가능한 볼륨이 선택된 경우 이 옵션은 생략 됩니다. 기본 설정 복원 복원 옵션은 모든 프로그램 매개변수를 기본 설정으로 복원합니다. 1. MENU(메뉴)를 누릅니다. 2. PROGRAM SETUP(프로그램 설정)>RESTORE(복원)로 이동합니다. 3. YES(예)를 선택합니다. 한글 169 프로그램 시작 또는 중지 RUN/HALT PROGRAM(프로그램 실행/중단) 키를 사용하여 프로그램을 시작하거나 중지합니다. 최대 12 개의 시작/중지 시간/날짜를 사용자 정의할 수 있습니다. 프로그램 또는 시스템 설정을 변경하려면 프로그램을 중지해야 합니다. 1. RUN/HALT PROGRAM(프로그램 실행/중단)을 누릅니다. 2. 옵션 중 하나를 선택합니다. 옵션 설명 시작 현재 로드된 프로그램을 시작합니다. 상태가 RUNNING(실행 중)으로 바뀝니다. 중단 프로그램을 일시적으로 중지합니다. 상태가 HALTED(중단됨)으로 바뀝니다. 재개 프로그램이 중단된 지점에서 재개됩니다. 상태가 RUNNING(실행 중)으로 바뀝니다. 처음부터 시작 프로그램이 처음부터 시작합니다. 상태가 RUNNING(실행 중)으로 바뀝니다. 프로그램 종료 현재 실행 중인 프로그램을 종료합니다. 상태가 COMPLETE(완료)로 바뀝니다. 수동 작동 수동 작동을 사용하여 그랩 샘플을 채취하거나 디스트리뷰터 암을 이동하거나 펌프를 작동합니다. 자 세한 내용은 본 설명서의 확장판을 참조하십시오. 볼륨 교정 샘플 볼륨은 액정 센서를 통해 자동으로, 또는 수동으로 교정할 수 있습니다. 액정 센서를 사용하여 볼 륨 교정을 실시할 경우 센서가 모든 프로그래밍된 볼륨을 정확하게 측정하도록 조정됩니다. 액정 센서 를 사용 안 함으로 설정한 경우에는 샘플, 린스 및 우수용 볼륨을 수동으로 교정해야 합니다. 일반 용도에서는 수돗물을 사용하여 볼륨 교정을 실시할 수 있습니다. 샘플 조성이 일반적인 물 샘플 과 현저히 다른 경우 샘플 채취기를 채취할 액체로 교정하십시오. 교정을 실시하려면 모든 프로그램을 중지해야 합니다. 액정 센서 교정 액정 센서의 정확도는 샘플로 채취되는 액체의 종류에 따라 다릅니다. 예를 들어 매우 혼탁한 샘플의 볼륨은 맑은 물만큼 정확하지 않을 수 있습니다. 액정 센서는 물 샘플로 교정하여 정확도를 개선할 수 있습니다. 1. 흡입 튜브를 물 샘플에 넣습니다. 2. MENU(메뉴)를 누릅니다. 3. PROGRAM SETUP(프로그램 설정)>MODIFY SELECTED(선택 수정)>LIQUID SENSOR(액정 센 서)>ENABLE(사용)으로 이동하여 액정 센서가 사용으로 설정되어 있는지 확인합니다. 4. SYSTEM SETUP(시스템 설정)>LIQ SENSOR CAL(액정 센서 교정)으로 이동합니다. 5. PERFORM CAL(교정 실시)을 선택합니다. 6. START(시작)를 선택합니다. 펌프가 역방향으로 작동하여 흡입 튜브를 정화합니다. 그런 다음 펌 프가 정방향으로 작동합니다. 7. 흡입 튜브 안의 샘플 흐름을 관찰하여 샘플이 액정 센서를 통과하는지 확인합니다. 8. 샘플이 액정 센서를 통과하고 기포가 없으면 STOP(중지)을 누릅니다. 흡입 튜브가 자동으로 정화 됩니다. 액정 센서 교정이 완료되었습니다. 9. DONE(완료)을 선택하여 종료하거나 REPEAT(반복)를 선택하여 교정을 다시 실시합니다. 기본 교정 복원 액정 센서의 볼륨 교정을 출고 시 기본 설정으로 복원할 수 있습니다. 170 한글 1. PROGRAM SETUP(프로그램 설정)>MODIFY SELECTED(선택 수정)>LIQUID SENSOR(액정 센 서)>ENABLE(사용)을 선택하여 액정 센서를 사용으로 설정합니다. 2. VOLUME CALIBRATION(볼륨 교정)을 누르고 RESET CAL(교정 재설정)을 선택합니다. 볼륨 교 정이 기본 설정으로 설정됩니다. “값이 0 으로 재설정됨” 메시지가 표시됩니다. 액정 센서를 사용하여 샘플 볼륨 교정 액정 센서를 사용으로 설정한 상태에서 샘플 볼륨을 교정하여 샘플 볼륨을 약간 조정합니다. 1. PROGRAM SETUP(프로그램 설정)>MODIFY SELECTED(선택 수정)>LIQUID SENSOR(액정 센 서)>ENABLE(사용)으로 이동하여 액정 센서가 사용으로 설정되어 있는지 확인합니다. 2. 흡입 튜브를 물 샘플 또는 수돗물에 넣습니다. 참고: 물 샘플을 사용하면 최상의 정확도를 얻을 수 있습니다. 3. 펌프의 출구 측에서 나온 튜브를 메스 실린더에 넣습니다. 4. VOLUME CALIBRATION(볼륨 교정)을 누르고 CALIBRATION(교정)을 선택합니다. 5. START(시작)를 선택합니다. 펌프가 역방향으로 작동하여 흡입 튜브를 정화합니다. 그런 다음 펌 프가 정방향으로 작동하여 샘플 채취 프로그램에 지정된 볼륨을 채취합니다. 펌프가 역방향으로 작동하여 흡입 튜브를 정화합니다. 6. 샘플이 채취되면 DONE(완료)을 선택합니다. 메스 실린더 안의 샘플 볼륨과 프로그램 설정 메뉴에 서 입력된 샘플 볼륨을 비교합니다. 7. 채취된 볼륨이 현재 프로그램의 샘플 볼륨과 다를 경우 실제로 채취된 볼륨을 입력합니다. ENTER 를 누릅니다. 센서가 모든 프로그래밍된 볼륨을 정확하게 측정하도록 조정됩니다. 8. START(시작)를 선택하여 샘플 채취 프로그램을 시작하거나 CANCEL(취소)을 선택하여 볼륨 교정 메뉴를 종료합니다. 시료 볼륨을 수동으로 교정 액정 센서를 사용 안 함으로 설정한 경우 모든 볼륨을 시간 기반으로 수동으로 교정해야 합니다. 샘플 볼륨이 현재 프로그램에 지정된 볼륨에 대해 교정됩니다. 프로그램에서 샘플 볼륨을 변경할 경우 새 볼륨에 대해 샘플 볼륨을 다시 교정해야 합니다. 1. PROGRAM SETUP(프로그램 설정)>MODIFY SELECTED(선택 수정)>LIQUID SENSOR(액정 센 서)>DISABLE(사용 안 함)을 선택하여 액정 센서를 사용 안 함으로 설정합니다. 2. 흡입 튜브를 물 샘플 또는 수돗물에 넣습니다. 3. 펌프의 출구 측에서 나온 튜브를 메스 실린더에 넣습니다. 4. VOLUME CALIBRATION(볼륨 교정)을 누릅니다. 2 개 이상의 볼륨이 표시되면 교정할 볼륨 1 개 를 선택합니다. 5. START(시작)를 선택합니다. 펌프가 역방향으로 작동하여 흡입 튜브를 정화합니다. 그런 다음 펌 프가 정방향으로 작동하여 샘플을 채취하기 시작합니다. 6. 선택된 볼륨 또는 린스 위치에서 펌프 작동을 중지합니다. • 샘플 볼륨: 메인 프로그램에서 지정된 볼륨이 채취되면 STOP(중지)을 누릅니다. • 린스: 흡입 튜브 안의 액체가 액체 검출기에 도달하면 STOP(중지)을 누릅니다. • 우수 볼륨: 우수 프로그램에서 지정된 볼륨이 채취되면 STOP(중지)을 누릅니다. 7. DONE(완료)을 선택하여 다음 볼륨 교정으로 이동하거나 REPEAT(반복)를 선택하여 교정을 다시 실시합니다. 8. 모든 볼륨을 교정했으면 DONE(완료)을 선택합니다. 9. 배출 튜브를 샘플 채취기의 튜브 피팅에 연결합니다. 10. START(시작)를 선택하여 샘플 채취 프로그램을 시작하거나 CANCEL(취소)을 선택하여 종료합니 다. 한글 171 샘플 볼륨 검증 샘플 볼륨이 정확한지 확인하려면 그랩 샘플을 채취합니다. 볼륨을 점검하기 위해 교정으로 돌아가지 마십시오. 교정 시작 시 볼륨 보상이 0 으로 재설정됩니다. 1. 2. 3. 4. 5. 6. 7. 8. MANUAL OPERATION(수동 작동)을 누릅니다. GRAB SAMPLE(그랩 샘플)을 선택합니다. 흡입 튜브를 물 샘플에 넣습니다. 펌프의 출구 측에서 나온 튜브를 메스 실린더에 넣습니다. 액정 센서를 사용으로 설정한 경우 검증할 볼륨을 입력합니다. 액정 센서를 사용 안 함으로 설정한 경우 프로그램에서 지정된 볼륨을 입력합니다. ENTER 를 누릅니다. 펌프 사이클이 시작합니다 (정화-샘플-정화). 메스 실린더에 채취된 볼륨을 예상 볼륨과 비교합니다. 채취된 볼륨이 정확하지 않으면 볼륨 교정 을 다시 실시합니다. 유지관리 위험 여러 가지 위험이 존재합니다. 해당 전문가가 본 문서에 의거하여 작업을 수행해야 합니다. 위험 전기쇼크 위험 유지관리 또는 정비 작업을 수행하기 전에 기기의 전원을 차단하십시오. 경고 생물학적 위험 노출. 샘플 병 및 샘플 채취기 성분과 접촉할 때는 안전 취급 규정을 따르십시오. 경고 여러 가지 위험이 존재합니다. 기술자는 유지관리 절차 후 장치가 안전하고 올바르게 작동하는지 확인해 야 합니다. 기기 세척 주의 화재 위험. 가연성 세정제를 사용하여 기기를 세척하지 마십시오. 주의사항 컨트롤러 히터를 어떠한 종류의 액체로도 청소하지 마십시오. 컨트롤러 및 펌프를 물로는 청소할 수 없을 경우 컨트롤러를 분리하여 샘플 채취기와 떨어뜨려 놓습니 다. 컨트롤러 및 펌프가 마르도록 충분히 기다린 후 부품을 다시 설치하고 사용을 재개하십시오. 다음과 같이 샘플 채취기를 청소합니다. • 샘플 채취기 케이스 - 내부 및 외부 표면은 젖은 헝겁과 중성 세제를 사용하여 청소하십시오. 마모성 클리너 또는 솔벤트를 사용하지 마십시오. • 샘플 병 - 병은 브러시와 중성 세제를 섞은 물을 사용하여 청소하십시오. 깨끗한 물로 헹군 다음 증 류수로 헹구십시오. 필요시 고압 멸균 처리기를 사용하여 유리병을 청소하십시오. 172 한글 건조제 교체 컨트롤러 내 건조제 카트리지는 습기를 흡수하고 부식을 방지합니다. 창을 통해 건조제 색상을 모니터 링합니다 (그림 12). 새 건조제는 노란색입니다. 녹색으로 변하면 건조제를 교체합니다. 1. 2. 3. 4. 5. 6. 나사를 풀고 건조제 카트리지를 분리합니다(그림 12). 플러그를 분리하고 사용한 건조제를 버립니다. 건조제 튜브에 새 건조제를 채웁니다. 플러그를 설치합니다. 오링에 그리스를 바릅니다. 건조제 튜브를 컨트롤러에 설치합니다. 그림 12 건조제 카트리지 1 플러그 3 오링 2 건조제 튜브 4 건조제 창 펌프 유지관리 주의 압착 위험. 유지관리 또는 정비 작업을 수행하기 전에 기기의 전원을 차단하십시오. 펌프 배관 교체 주의사항 제조업체에서 제공하지 않은 배관을 사용할 경우 기계 부품이 과도하게 마모될 수 있고 그리고/또는 펌프 성능이 저하될 수 있습니다. 롤러가 펌프 배관과 마찰되는 부위에서 배관의 마모 여부를 검사합니다. 알림 메시지가 표시되거나 배 관에서 마모 징후가 보이면 배관을 교체하십시오. 사전 필요 조건: • 펌프 배관—프리컷 또는 벌크 4.6 m 또는 15.2 m (15 ft 또는 50 ft) 1. 컨트롤러 측 전원을 분리합니다. 2. 벌크 배관을 사용하는 경우 배관을 절단하고 정렬 도트를 추가합니다. 제조업체 웹 사이트의 세부 설명서를 참조하십시오. 3. 아래의 그림 단계에 따라 펌프 배관을 분리합니다. 4. 펌프 하우징 내부와 롤러에서 실리콘 잔류물을 제거합니다. 5. 아래의 그림 단계에 따라 펌프 배관을 설치합니다. 한글 173 174 한글 로터 청소 연성 세제로 로터, 펌프 튜브 트랙 및 펌프 하우징을 청소합니다. 펌프 배관 교체 페이지의 173 와 아래 의 그림 단계를 참조하십시오. 디스트리뷰터 암 튜브 교체 디스트리뷰터 암은 다중 용기 샘플 채취 시 각 병 위로 이동합니다. 튜브가 마모된 경우 디스트리뷰터 암의 튜브를 교체합니다. 디스트리뷰터 및 디스트리뷰터 암에 적합한 튜브를 사용해야 합니다. 참고: 디스트리뷰터와 펌프 튜브는 서로 다릅니다. 디스트리뷰터 어셈블리에 설치된 펌프 튜브로 인해 디스트리뷰터 가 손상될 수 있습니다. 또한 디스트리뷰터 암이 제대로 이동하지 못하여 샘플이 손실될 수 있습니다. 1. 2. 3. 4. 디스트리뷰터 암과 의 중심부 천정에서 튜브를 제거합니다. 디스트리뷰터 암에 새 튜브를 끼웁니다. 튜브는 그림 13 과 같이 규격에 맞춰야 합니다. 튜브의 다른 쪽 끝을 의 중심부 천정에 있는 피팅에 끼웁니다. 디스트리뷰터 진단 시험을 실시하여 올바로 작동하는지 확인합니다. 그림 13 디스트리뷰터 어셈블리 1 노즐 3 디스트리뷰터 모터 2 디스트리뷰터 암 4 샤프트 5 디스트리뷰터 튜브 퓨즈 교체 퓨즈는 사용자가 취급할 수 있는 품목이 아닙니다. 기술 지원부에 문의하십시오. 퓨즈 단절은 샘플 채 취기에 문제가 생겼다는 징후이며, 이 경우에는 서비스를 요청해야 합니다. 한글 175 HACH COMPANY World Headquarters P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 [email protected] www.hach.com © HACH LANGE GMBH Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 [email protected] www.hach-lange.de Hach Company/Hach Lange GmbH, 2013. All rights reserved. Printed in U.S.A. HACH LANGE Sàrl 6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178

Hach SD900 Basic User Manual

Tipo
Basic User Manual

en otros idiomas