KS4.3 T
0.1 - 4 mmol/l KS4.3
Acid / Indicator
20
S:4.3
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette
and the absorption range of the photometer are indicated.
Instrument Type Cuvette λMeasuring Range
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Alka-M-Photometer Tablet / 100 513210BT
Alka-M-Photometer Tablet / 250 513211BT
Application List
Waste Water Treatment
Drinking Water Treatment
Raw Water Treatment
Notes
1. The terms Alkalinity-m, m-Value, total alkalinity and Acid demand to KS4.3 are iden-
tical.
2. For accurate results, exactly 10 ml of water sample must be used for the test.
Display in the MD
100 / MD 110 /
MD 200
Bar code for the
detection of the
methods
Method name
Measuring range
Chemical Method
Method number
Language codes
ISO 639-1 Revision status
KS4.3 T / 20
How to use
ENMethod Reference Book4
EN
Implementation of the provision Acid capacity KS4.3 with Tablet
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
10 ml
Fill 24 mm vial with 10 ml
sample.
Close vial(s).
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Dissolve tablet(s) by inver-
ting.
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Test
Press the TEST (XD:
START) button.
The result in Acid Capacity KS4.3 appears on the display.
Performing test
procedure
KS4.3 T / 20
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
How to use
ENMethod Reference Book 5
EN
Suspended solids 24 / M384
ENMethod Reference Book1.06
EN
Suspended solids 24 M384
10 - 750 mg/L TSS SuS
Turbidity / Attenuated Radiation Method
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
no reagent required
Sampling
1. Measure the water sample as soon as possible after sampling. It is possible to store
the sample at 4 °C for 7 days s in plastic or glass containers. The measurement
should be at the same temperature as the sample. Temperature differences
between measurement and sampling can change the result of the measurement.
Remarks
1. The photometric determination of Suspended Solids is based on a gravimetric
method. In a laboratory this is usually done by evaporation of the filter residue of
a filtrated water sample in a furnace at 103 °C – 105 °C and weighing of the dried
residue.
2. When higher accuracy is required perform a gravimetric determination of a water
sample. The result can be used to calibrate the photometer with the same water
sample.
3. The estimated detection limit is 20 mg/L TSS.
Suspended solids 24 / M384
ENMethod Reference Book1.0 7
EN
Determination of Total suspended solids
Select the method on the device.
Homogenize mL of the water sample in a blender on high speed for minutes
10 mL
Sample
Fill 24 mm vial with 10 mL
deionised water .
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
24 mm
24 mm
10 mL
Mix homogenised water
sample thoroughly.
Pre-rinse vial with water
sample.
Fill 24 mm vial with 10 mL
prepared sample .
Suspended solids 24 / M384
ENMethod Reference Book1.08
EN
Sample
Test
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
The result in mg/L TSS (Total Suspended Solids) appears on the display.
Suspended solids 24 / M384
ENMethod Reference Book1.0 9
EN
Chemical Method
Turbidity / Attenuated Radiation Method
Appendix
Interferences
Persistant Interferences
Colour interferes if light is absorbed at 660 nm.
Removeable Interferences
Air bubbles interfere and can be removed by swirling the vial gently.
Method Validation
Limit of Detection 10 mg/L
Limit of Quantification 30 mg/L
End of Measuring Range 750 mg/L
Sensitivity 550 mg/L / Abs
Confidence Intervall 4.24 mg/L
Standard Deviation 1.79 mg/L
Variation Coefficient 0.47 %
Derived from
EN 872:2005
DE Methodenhandbuch 01/20
KS4.3 T
0,1 - 4 mmol/l KS4.3
Säure / Indikator
20
S:4.3
Instrumentspezifische Informationen
Der Test kann auf den folgenden Geräten durchgeführt werden. Zusätzlich sind die
benötigte Küvette und der Absorptionsbereich der Photometer angegeben.
Geräte Küvette λMessbereich
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0,1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0,1 - 4 mmol/l KS4.3
Material
Benötigtes Material (zum Teil optional):
Reagenzien Form/Menge Bestell-Nr.
Alka-M-Photometer Tablette / 100 513210BT
Alka-M-Photometer Tablette / 250 513211BT
Anwendungsbereich
Abwasserbehandlung
Trinkwasseraufbereitung
Rohwasserbehandlung
Anmerkungen
1. Die Begrie Alkalität-m, m-Wert, Gesamtalkalität und Säurekapazität KS4.3 sind
identisch.
2. Die exakte Einhaltung des Probevolumens von 10 ml ist für die Genauigkeit des
Analysenergebnisses entscheidend.
Displayanzeige
im MD 100 MD 110
/ MD 200
Barcode zur
Methodenerkennung
Methoden Name
Messbereich
Chemische
Methode
Methodennummer
Sprachkürzel nach
ISO 639-1 Revisionsstand
KS4.3 T / 20
Leitfaden für den Anwender
ENMethod Reference Book10
DE
Durchführung der Bestimmung Säurekapazität KS4.3 mit Tablette
Die Methode im Gerät auswählen.
Für diese Methode muss bei folgenden Geräten keine ZERO-Messung durchgeführt
werden: XD 7000, XD 7500
10 ml
24-mm-Küvette mit 10 ml
Probe füllen.
Küvette(n) verschließen.
Sample
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Tablette(n) durch Um-
schwenken lösen.
Sample
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Test
Taste TEST (XD: START)
drücken.
In der Anzeige erscheint das Ergebnis als Säurekapazität KS4.3.
Durchführung der
Messung
KS4.3 T / 20
DE Methodenhandbuch 01/20
Leitfaden für den Anwender
ENMethod Reference Book 11
DE
Suspend. Feststoffe 24 / M384
DEMethod Reference Book1.012
DE
Suspend. Feststoffe 24 M384
10 - 750 mg/L TSS SuS
Trübung / Durchlicht
Material
Benötigtes Material (zum Teil optional):
Reagenzien Form/Menge Bestell-Nr.
kein Reagenz erforderlich
Probenahme
1. Die Wasserprobe so schnell wie möglich nach der Probenahme messen. Proben
können bis zu 7 Tage bei 4 °C in Kunststoff- oder Glasflaschen aufbewahrt werden.
Die Messung sollte bei derselben Temperatur erfolgen, wie bei Entnahme der
Probe. Temperaturunterschiede zwischen Messung und Probenahme können das
Messergebnis verändern.
Anmerkungen
1. Die photometrische Bestimmung der suspendierten Feststoffe basiert auf
einer gravimetrischen Methode. In einem Labor wird das Eindampfen des
Filterrückstandes einer abfiltrierten Wasserprobe gewöhnlich in einem Ofen bei 103
°C - 105 °C vorgenommen und der getrocknete Rückstand ausgewogen.
2. Wird eine erhöhte Genauigkeit benötigt, so ist eine gravimetrische Bestimmung
einer Probe durchzuführen. Dieses Ergebnis kann für eine Anwender-Justierung
des Photometers mit derselben Probe verwendet werden.
3. Die geschätzte Nachweisgrenze für diese Methode liegt bei 20 mg/L TSS.
Suspend. Feststoffe 24 / M384
DEMethod Reference Book1.0 13
DE
Durchführung der Bestimmung Suspendierte Feststoffe
Die Methode im Gerät auswählen.
mL der Wasserprobe in einem Mixer auf hoher Stufe für Minuten homogenisieren.
10 mL
Sample
24 mm-Küvette mit 10 mL
VE-Wasser füllen.
Küvette(n) verschließen. Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Zero
Taste ZERO drücken. Küvette aus dem
Messschacht nehmen.
24 mm
24 mm
10 mL
Homogenisierte
Wasserprobe gut
durchmischen.
Die Küvette mit der
Wasserprobe vorspülen.
24-mm-Küvette mit 10 mL
vorbereiteter Probe füllen.
Suspend. Feststoffe 24 / M384
DEMethod Reference Book1.014
DE
Sample
Test
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Taste TEST (XD: START)
drücken.
In der Anzeige erscheint das Ergebnis in mg/L TSS (Total Suspended Solids).
Suspend. Feststoffe 24 / M384
DEMethod Reference Book1.0 15
DE
Chemische Methode
Trübung / Durchlicht
Appendix
Störungen
Permanente Störungen
Farbe stört, wenn Licht bei 660 nm absorbiert wird.
Ausschließbare Störungen
Luftblasen stören und können durch leichtes Schwenken der Küvette entfernt
werden.
Methodenvalidierung
Nachweisgrenze 10 mg/L
Bestimmungsgrenze 30 mg/L
Messbereichsende 750 mg/L
Empfindlichkeit 550 mg/L / Abs
Vertrauensbereich 4.24 mg/L
Verfahrensstandardabweichung 1.79 mg/L
Verfahrensvariationskoeffizient 0.47 %
Abgeleitet von
EN 872:2005
DE Methodenhandbuch 01/20ES Manual de Métodos 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Ácido / Indicador
20
S:4.3
Información específica del instrumento
La prueba puede realizarse en los siguientes dispositivos. Además, se muestran la
cubeta requerida y el rango de absorción del fotómetro.
Dispositivos Cubeta λRango de medición
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Material
Material requerido (parcialmente opcional):
Título Unidad de
embalaje
Referencia
No
Fotómetro alca-M Tabletas / 100 513210BT
Fotómetro alca-M Tabletas / 250 513211BT
Lista de aplicaciones
Tratamiento de aguas residuales
Tratamiento de aguas potables
Tratamiento de aguas de aporte
Notas
1. Las deniciones de alcalinidad-m, valor-m y capacidad ácida KS4.3 son idénticas.
2. Añadir un volumen de muestra de exactamente 10 ml, ya que este volumen inuye
de forma decisiva en la exactitud del resultado.
Indicación en la
pantalla de MD 100
/ MD 110 / MD 200
Código de barras
para reconocer el
método
Nombre del
método
Rango de
medición
Método químico
Número de
método
Códigos de idioma
ISO 639-1 Estado de revisión
KS4.3 T / 20
How to use
ENMethod Reference Book16
ES
Ejecución de la determinación Capacidad ácida KS4.3 con tableta
Seleccionar el método en el aparato.
Para este método no es necesario realizar medición CERO en los aparatos siguientes:
XD 7000, XD 7500
10 ml
Llenar la cubeta de 24 mm
con 10 ml de muestra .
Cerrar la(s) cubeta(s).
Sample
Poner la cubeta de muest-
ra en el compartimiento de
medición. ¡Debe tenerse en
cuenta el posicionamiento!
Añadir tableta ALKA-M-
PHOTOMETER.
24 mm
Triturar la(s) tableta(s) giran-
do ligeramente.
Cerrar la(s) cubeta(s).
Realización de la
determinación
KS4.3 T / 20
DE Methodenhandbuch 01/20ES Manual de Métodos 01/20
How to use
ENMethod Reference Book 17
ES
Sustancias sólidas suspend. 24 / M384
ESMethod Reference Book1.018
ES
Sustancias sólidas suspend. 24 M384
10 - 750 mg/L TSS SuS
Turbidez / método de radiación
atenuada
Material
Material requerido (parcialmente opcional):
Reactivos Unidad de
embalaje
No. de
referencia
sin necesidad de reactivo
Muestreo
1. Medir la muestra acuosa lo antes posible después de la toma de la muestra. Las
muestras se pueden guardar hasta 7 días a 4 °C en botellas de plástico o de vidrio.
La medición se deberá realizar a la misma temperatura a la que se ha tomado la
muestra. Las diferencias de temperatura entre la medición y la toma de la muestra
pueden modificar el resultado de la medición.
Notas
1. La determinación fotométrica de la sustancia sólida suspendida se basa en
un método gravimétrico. En un laboratorio se lleva a cabo habitualmente la
evaporación de los residuos del filtro de una muestra de agua filtrada en un horno a
103 °C - 105 °C y se pesa el residuo seco.
2. Si se necesita una exactitud mayor, se deberá realizar una determinación
gravimétrica de una muestra. Este resultado puede ser utilizado con la misma
muestra para un ajuste de usuario del fotómetro.
3. El límite de determinación calculado para este método es de aprox. 20 mg/L TSS.
Sustancias sólidas suspend. 24 / M384
ESMethod Reference Book1.0 19
ES
Ejecución de la determinación Sustancias sólidas suspendidas
Seleccionar el método en el aparato.
Homogeneizar mL de muestra de agua en un batidor, en la potencia más alta durante
minutos.
10 mL
Sample
Llenar la cubeta de 24 mm
con 10 mL de agua
desionizada.
Cerrar la(s) cubeta(s). Poner la cubeta de muestra
en el compartimiento de
medición. ¡Debe tenerse en
cuenta el posicionamiento!
Zero
Pulsar la tecla ZERO. Extraer la cubeta del
compartimiento de
medición.
24 mm
24 mm
10 mL
Mezclar bien la muestra
acuosa homogeneizada.
Prelavar la cubeta con la
muestra acuosa.
Llenar la cubeta de 24 mm
con 10 mL de la muestra
preparada .
Sustancias sólidas suspend. 24 / M384
ESMethod Reference Book1.020
ES
Sample
Test
Poner la cubeta
de muestra en el
compartimiento de
medición. ¡Debe tenerse en
cuenta el posicionamiento!
Pulsar la tecla TEST (XD:
START).
A continuación se visualizará el resultado en mg/L TSS (Total Sólidos Disueltos).
Sustancias sólidas suspend. 24 / M384
ESMethod Reference Book1.0 21
ES
Método químico
Turbidez / método de radiación atenuada
Apéndice
Interferencia
Interferencias persistentes
Si la luz se absorbe a 660 nm se perturba el color.
Interferencias extraibles
Las burbujas de aire alteran y pueden ser eliminadas haciendo girar suavemente la
cubeta.
Validación del método
Límite de detección 10 mg/L
Límite de determinación 30 mg/L
Límite del rango de medición 750 mg/L
Sensibilidad 550 mg/L / Abs
Intervalo de confianza 4.24 mg/L
Desviación estándar 1.79 mg/L
Coeficiente de variación 0.47 %
Derivado de
EN 872:2005
DE Methodenhandbuch 01/20FR Méthodes Manuel 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Acide / Indicateur
20
S:4.3
Informations spécifiques à l‘instrument
Le test peut être eectué sur les appareils suivants. De plus, la cuvette requise et la
plage d‘absorption du photomètre sont indiquées.
Appareils Cuvette λGamme de mesure
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Matériel
Matériel requis (partiellement optionnel):
Titre Pack contenant Code
Alka-M-Photometer Pastilles / 100 513210BT
Alka-M-Photometer Pastilles / 250 513211BT
Liste d‘applications
Traitement des eaux usées
Traitement de l’eau potable
Traitement de l’eau brute
Indication
1. Les termes Alcalinité-m, Valeur m, Alcalinité totale et Capacité acide KS4.3 sont
identiques.
2. L’observation exacte du volume d’échantillon de 10 ml est décisive pour l‘exactitude
du résultat de l’analyse.
Achage dans le
MD 100 / MD 110 /
MD 200
Code à barres
pour reconnaître la
méthode
Nom de la
méthode
Plage de mesure
Méthode chimique
Numéro de
méthode
Codes de langue
ISO 639-1 État de révision
KS4.3 T / 20
How to use
ENMethod Reference Book22
FR
DE Methodenhandbuch 01/20FR Méthodes Manuel 01/20
Réalisation de la quantification Capacité acide KS4.3 avec pastille
Sélectionnez la méthode sur l’appareil.
Cette méthode ne nécessite aucune mesure du zéro sur les appareils suivants : XD
7000, XD 7500
10 ml
Remplissez une cuvet-
te de 24 mm de 10 ml
d’échantillon.
Fermez la(les) cuvette(s).
Sample
Placez la cuvette réser-
vée à l’échantillon dans
la chambre de mesure.
Attention à la positionner
correctement.
Ajoutez une pastille de
ALKA-M-PHOTOMETER.
24 mm
Écrasez la(les) pastille(s) en
la(les) tournant un peu.
Fermez la(les) cuvette(s).
Procédure du test
KS4.3 T / 20
How to use
ENMethod Reference Book 23
FR
Solides en suspension 24 / M384
FRMethod Reference Book1.024
FR
Solides en suspension 24 M384
10 - 750 mg/L TSS SuS
Turbidité/méthode de radiation atténuée
Matériel
Matériel requis (partiellement optionnel):
Réactifs Pack contenant Code
Ne nécessite aucun réactif
Échantillonnage
1. Mesurez l’échantillon d'eau le plus rapidement possible après le prélèvement. Les
échantillons peuvent être conservés pendant 7 jours max. à une température de 4
°C dans des bouteilles en plastique ou en verre. La mesure devrait avoir lieu à la
même température que celle du prélèvement de l’échantillon. Les différences de
température entre la mesure et le prélèvement peuvent modifier le résultat.
Indication
1. La quantification photométrique des solides en suspension est basée sur une
méthode gravimétrique. En laboratoire, l'évaporation du résidu de filtration d'un
échantillon d'eau filtré a généralement lieu dans un four à 103 °C - 105 °C. Le
résidu séché est alors pesé.
2. Si une exactitude supérieure est nécessaire, on procède à une quantification
gravimétrique d’un échantillon. Ce résultat peut être utilisé pour ajuster le
photomètre avec le même échantillon.
3. La limite de détection estimée pour cette méthode est de 20 mg/L TSS.
Solides en suspension 24 / M384
FRMethod Reference Book1.0 25
FR
Réalisation de la quantification Solides en suspension
Sélectionnez la méthode sur l’appareil.
Homogénéisez mL de l’échantillon d'eau dans un agitateur, à grande vitesse et pendant
minutes.
10 mL
Sample
Remplissez une cuvette de
24 mm de 10 mL d’eau
déminéralisée.
Fermez la(les) cuvette(s). Placez la cuvette réservée
à l’échantillon dans la
chambre de mesure.
Attention à la positionner
correctement.
Zero
Appuyez sur la touche
ZERO.
Retirez la cuvette de la
chambre de mesure.
24 mm
24 mm
10 mL
Mélangez correctement
l’échantillon d'eau
homogénéisée.
Prélavez la cuvette avec
l’échantillon d'eau.
Remplissez une cuvette
de 24 mm de 10 mL
d’échantillon préparé.
Solides en suspension 24 / M384
FRMethod Reference Book1.026
FR
Sample
Test
Placez la cuvette réservée
à l’échantillon dans la
chambre de mesure.
Attention à la positionner
correctement.
Appuyez sur la touche
TEST (XD: START).
Le résultat s'affiche à l’écran en mg/L TSS (Matières en suspension totales).
Solides en suspension 24 / M384
FRMethod Reference Book1.0 27
FR
Méthode chimique
Turbidité/méthode de radiation atténuée
Appendice
Interférences
Interférences persistantes
La couleur perturbe si la lumière est absorbée à 660 nm.
Interférences exclues
Les bulles d'air perturbent la mesure et sont éliminées en agitant légèrement la
cuvette.
Méthode Validation
Limite de détection 10 mg/L
Limite de détermination 30 mg/L
Fin de la gamme de mesure 750 mg/L
Sensibilité 550 mg/L / Abs
Intervalle de confiance 4.24 mg/L
Déviation standard 1.79 mg/L
Coefficient de variation 0.47 %
Dérivé de
EN 872:2005
DE Methodenhandbuch 01/20IT Manuale dei Metodi 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Acido/indicatore
20
S:4.3
Informazioni specifiche dello strumento
Il test può essere eseguito sui seguenti dispositivi. Inoltre, sono indicate la cuvetta
richiesta e il range di assorbimento del fotometro.
Dispositivi Cuvetta λCampo di misura
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Materiale
Materiale richiesto (in parte facoltativo):
Titolo Unità di
imballaggio
N. ordine
Alka-M-Photometer Pastiglia / 100 513210BT
Alka-M-Photometer Pastiglia / 250 513211BT
Campo di applicazione
Trattamento acqua di scarico
Trattamento acqua potabile
Trattamento acqua non depurata
Note
1. I termini alcalinità M, valore M, alcalinità totale e capacità acida KS4.3 sono equiva-
lenti.
2. Per l’accuratezza del risultato dell’analisi è fondamentale che il volume del campio-
ne misuri esattamente 10 ml.
Indicazione sul
display del MD 100
/ MD 110 / MD 200
Codice a barre
per riconoscere il
metodo
Denominazione
metodo
Range di misura
Metodo chimico
Numero metodo
ISO 639-1 codici
linguistici Stato di revisione
KS4.3 T / 20
How to use
ENMethod Reference Book28
IT
Esecuzione della rilevazione Capacità acida KS4.3 con pastiglia
Selezionare il metodo nel dispositivo.
Con i seguenti dispositivi, per questo metodo non è necessario eseguire una misurazio-
ne ZERO: XD 7000, XD 7500
10 ml
Riempire una cuvetta da 24
mm con 10 ml di campione.
Chiudere la/e cuvetta/e.
Sample
Posizionare la cuvetta del
campione nel vano di misu-
razione. Fare attenzione al
posizionamento.
Aggiungere una pastiglia
ALKA-M-PHOTOMETER.
24 mm
Frantumare la/e pastiglia/e
con una leggera rotazione.
Chiudere la/e cuvetta/e.
Svolgimento della
misurazione
KS4.3 T / 20
DE Methodenhandbuch 01/20IT Manuale dei Metodi 01/20
How to use
ENMethod Reference Book 29
IT
Solidi sospesi 24 / M384
ITMethod Reference Book1.030
IT
Solidi sospesi 24 M384
10 - 750 mg/L TSS SuS
Torbidità / luce trasmessa
Materiale
Materiale richiesto (in parte facoltativo):
Reagenti Unità di
imballaggio
N. ordine
Nessun reagente richiesto
Prelievo del campione
1. Il campione di acqua deve essere misurato al più presto dopo il prelievo. I campioni
possono essere conservati fino a 7 giorni a 4 °C in flaconi di plastica o vetro. La
misurazione dovrebbe avvenire alla stessa temperatura presente al momento del
prelievo del campione. Eventuali differenze di temperatura tra la misurazione e il
prelievo del campione possono modificare il risultato della misurazione.
Note
1. La determinazione fotometrica dei solidi sospesi è basata su un metodo
gravimetrico. In un laboratorio la vaporizzazione del residuo di filtrazione di un
campione di acqua filtrato viene solitamente eseguita in un forno a 103-105 °C e il
residuo essiccato viene bilanciato.
2. Se è richiesta un’accuratezza elevata è necessario eseguire una determinazione
gravimetrica di un campione. Questo risultato può essere utilizzato per una
regolazione personalizzata del fotometro con lo stesso campione.
3. Il limite di rilevabilità stimato per questo metodo è di 20 mg/L di TSS.
Solidi sospesi 24 / M384
ITMethod Reference Book1.0 31
IT
Esecuzione della rilevazione Solidi sospesi
Selezionare il metodo nel dispositivo.
Omogeneizzare mL del campione di acqua in un agitatore a velocità elevata per minuti.
10 mL
Sample
Riempire una cuvetta da
24 mm con 10 mL di
acqua demineralizzata.
Chiudere la/e cuvetta/e. Posizionare la cuvetta
del campione nel vano di
misurazione. Fare attenzione
al posizionamento.
Zero
Premere il tasto ZERO. Prelevare la cuvetta dal
vano di misurazione.
24 mm
24 mm
10 mL
Miscelare bene il campione
di acqua omogeneizzato.
Sciacquare
preventivamente la cuvetta
con il campione di acqua.
Riempire una cuvetta da
24 mm con 10 mL del
campione preparato.
Solidi sospesi 24 / M384
ITMethod Reference Book1.032
IT
Sample
Test
Posizionare la cuvetta
del campione nel
vano di misurazione.
Fare attenzione al
posizionamento.
Premere il tasto TEST (XD:
START).
Sul display compare il risultato in mg/L di TSS (Totale solidi in sospensione).
Solidi sospesi 24 / M384
ITMethod Reference Book1.0 33
IT
Metodo chimico
Torbidità / luce trasmessa
Appendice
Interferenze
Interferenze permanenti
Il colore provoca interferenze se la luce viene assorbita a 660 nm.
Interferenze escludibili
Le bolle d'aria provocano interferenze e possono essere rimosse facendo oscillare
leggermente la cuvetta.
Validazione metodo
Limite di rilevabilità 10 mg/L
Limite di quantificazione 30 mg/L
Estremità campo di misura 750 mg/L
Sensibilità 550 mg/L / Abs
Intervallo di confidenza 4.24 mg/L
Deviazione standard della
procedura
1.79 mg/L
Coefficiente di variazione della
procedura
0.47 %
Derivato di
EN 872:2005
DE Methodenhandbuch 01/20PT Métodos Manual 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Ácido / Indicador
20
S:4.3
Informação específica do instrumento
O teste pode ser realizado nos seguintes dispositivos. Além disso, a cubeta necessária
e a faixa de absorção do fotômetro são indicadas.
Dispositivos Cubeta λFaixa de Medição
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Material
Material necessário (parcialmente opcional):
Título Unidade de
Embalagem
Artigo No
Alka-M-Photometer Pastilhas / 100 513210BT
Alka-M-Photometer Pastilhas / 250 513211BT
Lista de Aplicações
Tratamento de Esgotos
Tratamento de Água Potável
Tratamento de Água Bruta
Notas
1. Os termos alcalinidade-m, m-valor, alcalinidade total e capacidade de acidez KS4.3
são idênticos.
2. O cumprimento exato do volume da amostra de 10 ml é decisivo para a precisão do
resultado de análise.
Indicado no
display: MD 100
MD 110 / MD 200
Código de barras
para a detecção dos
métodos
Nome do método
Área de medição
Método Químico
Número do
método
Códigos de idioma
ISO 639-1 Nível de revisão
KS4.3 T / 20
How to use
ENMethod Reference Book34
PT
DE Methodenhandbuch 01/20PT Métodos Manual 01/20
Realização da determinação Capacidade de acidez KS4.3 com
pastilha
Escolher o método no equipamento.
Para este método não tem de ser efetuada uma medição ZERO nos seguintes equipa-
mentos: XD 7000, XD 7500
10 ml
Encher a célula de 24 mm
com 10 ml de amostra .
Fechar a(s) célula(s).
Sample
Colocar a célula de
amostra no compartimento
de medição. Observar o
posicionamento.
Pastilha ALKA-M-PHOTO-
METER.
24 mm
Esmagar a(s) pastilha(s)
rodando ligeiramente.
Fechar a(s) célula(s).
Efetuar a medição
KS4.3 T / 20
How to use
ENMethod Reference Book 35
PT
Matéria sólida suspensa 24 / M384
PTMethod Reference Book1.036
PT
Matéria sólida suspensa 24 M384
10 - 750 mg/L TSS SuS
Turbidez / Método de Radiação
Atenuada
Material
Material necessário (parcialmente opcional):
Reagentes Unidade de
Embalagem
Código do
Produto
não é necessário reagente
Amostragem
1. Medir a amostra de água logo após a recolha da amostra. As amostras podem
ser guardadas até 7 dias a 4 °C em garrafas de plástico ou de vidro. A medição
devia ser efetuada à mesma temperatura da recolha da amostra. As diferenças de
temperatura entre a medição e a recolha da amostra podem alterar o resultado de
medição.
Notas
1. A determinação fotométrica da matéria sólida suspensa baseia-se num método
gravimétrico. Num laboratório procede-se à evaporação do resíduo de filtragem
de uma amostra de água filtrada normalmente num forno a 103 °C - 105 °C, e o
resíduo seco é equilibrado.
2. Se precisar de mais precisão, deve realizar uma determinação gravimétrica de uma
amostra. Este resultado pode ser usado para um ajuste do utilizador do fotómetro
com a mesma amostra.
3. O limite de prova estimado para este método situa-se em 20 mg/L TSS.
Matéria sólida suspensa 24 / M384
PTMethod Reference Book1.0 37
PT
Realização da determinação Matéria sólida suspensa
Escolher o método no equipamento.
Homogeneizar mL da amostra de água num misturador a alto nível durante minutos.
10 mL
Sample
Encher a célula de 24 mm
com 10 mL de água
desmineralizada.
Fechar a(s) célula(s). Colocar a célula de
amostra no compartimento
de medição. Observar o
posicionamento.
Zero
Premir a tecla ZERO. Retirar a célula do
compartimento de
medição.
24 mm
24 mm
10 mL
Misturar bem a amostra de
água homogeneizada.
Pré-enxaguar a célula com
a amostra de água.
Encher a célula de 24 mm
com 10 mL de amostra
preparada.
Matéria sólida suspensa 24 / M384
PTMethod Reference Book1.038
PT
Sample
Test
Colocar a célula de
amostra no compartimento
de medição. Observar o
posicionamento.
Premir a tecla TEST (XD:
START).
No visor aparece o resultado em mg/L TSS (Total de Sólidos Suspensos).
Matéria sólida suspensa 24 / M384
PTMethod Reference Book1.0 39
PT
Método Químico
Turbidez / Método de Radiação Atenuada
Apêndice
Texto de Interferências
Interferências Pesistentes
A cor interfere quando a luz é absorvida a 660 nm.
Interferências Removíveis
As bolhas de ar interferem e podem ser removidas se agitar ligeiramente a célula.
Validação de método
Limite de Detecção 10 mg/L
Limite de Determinação 30 mg/L
Fim da Faixa de Medição 750 mg/L
Sensibilidade 550 mg/L / Abs
Faixa de Confiança 4.24 mg/L
Desvio Padrão 1.79 mg/L
Coeficiente de Variação 0.47 %
Derivado de
EN 872:2005
NL Handboek 'Methoden'01/20NL Handboek van Methoden 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Zuur / Indicator
M20
S:4.3
Instrumentspecifieke informatie
De test kan op de volgende apparaten worden uitgevoerd. Bovendien worden de vereis-
te cuvette en het absorptiebereik van de fotometer aangegeven.
Toestellen Cuvet λMeetbereik
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Reagentia
Benodigd materiaal (deels optioneel):
Titel Verpakkingseenheid Bestelnr.
Alka-M-Photometer Tablet / 100 513210BT
Alka-M-Photometer Tablet / 250 513211BT
Toepassingsbereik
Afvalwaterzuivering
Behandeling drinkwater
Zuivering vervuild water
Aantekeningen
1. De termen alkaliteit-m, m-waarde, totale alkaliteit en zuurcapaciteitKS4.3 zijn identiek.
2. De exacte naleving van het monstervolume van 10 ml is bepalend voor de
nauwkeurigheid van het analyseresultaat.
Uitlezing in MD
100 MD 110 / MD
200
Streepjescode ter
identicatie van de
methode
Naam van de
methode
Meetbereik
Chemische
methode
Nummer methode
Beknopte naam
conform de norm
ISO 639-1
Herziene versie
KS4.3 T / 20
How to use
ENMethod Reference Book40
NL
NL Handboek 'Methoden'01/20NL Handboek van Methoden 01/20
Uitvoering van de bepaling Zuurcapaciteit KS4.3 met tablet
De methode in het apparaat selecteren.
Voor deze methode moet bij de volgende apparaten geen nulmeting worden uitgevoerd:
XD 7000, XD 7500
10 ml
Spoelbakje van 24 mm
met 10 ml staal vullen.
De spoelbakjes afsluiten.
Sample
Het staalspoelbakje in de
meetschacht plaatsen. Op
de positionering letten.
Tabletten oplossen door om
te draaien
Sample
Het staalspoelbakje in de
meetschacht plaatsen. Op
de positionering letten.
Test
De toets TEST (XD: START)
indrukken.
De display toont het resultaat als Zuurcapaciteit KS4.3.
Uitvoering van de
meting
KS4.3 T / 20
How to use
ENMethod Reference Book 41
NL
Zwevende Vaste stoffen 24 / M384
NLMethod Reference Book1.042
NL
Zwevende Vaste stoffen 24 M384
10 - 750 mg/L TSS SuS
Troebelheid / doorvallend licht
Reagentia
Benodigd materiaal (deels optioneel):
Reagentia VerpakkingseenheidBestelnr.
geen reagens nodig
Bemonstering
1. Meet het watermonster zo spoedig mogelijk na de bemonstering. Monsters kunnen
tot 7 dagen bij 4 °C worden bewaard in plastic of in glazen flessen. De meting moet
worden uitgevoerd bij dezelfde temperatuur als toen het monster werd genomen.
Temperatuurverschillen tussen meting en bemonstering kunnen het meetresultaat
beïnvloeden.
Aantekeningen
1. De fotometrische bepaling van de zwevende vaste stoffen is gebaseerd op
een gravimetrische methode. In een laboratorium wordt het filterresidu van een
gefiltreerd watermonster meestal verdampt in een oven bij 103 - 105°C en het
gedroogde residu wordt gebalanceerd.
2. Indien een grotere nauwkeurigheid vereist is, moet een gravimetrische bepaling
van een monster worden uitgevoerd. Dit resultaat kan worden gebruikt als
gebruikersverantwoording van de fotometer met hetzelfde monster.
3. De geschatte aantoonbaarheidsgrens voor deze methode is 20 mg/L TSS.
Zwevende Vaste stoffen 24 / M384
NLMethod Reference Book1.0 43
NL
Uitvoering van de bepaling Zwevende vaste stoffen
De methode in het apparaat selecteren.
mL van het waterstaal gedurende minuten in een mixer op hoge stand homogeniseren.
10 mL
Sample
Spoelbakje van 24 mm
met 10 mL gedeïoniseerd
water vullen.
De spoelbakjes afsluiten. Het staalspoelbakje in de
meetschacht plaatsen. Op
de positionering letten.
Zero
De toets NUL indrukken. Het spoelbakje uit de
meetschacht nemen.
24 mm
24 mm
10 mL
Het gehomogeniseerde
waterstaal goed mengen.
Het spoelbakje met het
waterstaal voorspoelen.
Spoelbakje van 24 mm met
10 mL voorbereid staal
vullen.
Zwevende Vaste stoffen 24 / M384
NLMethod Reference Book1.044
NL
Sample
Test
Het staalspoelbakje in de
meetschacht plaatsen. Op
de positionering letten.
De toets TEST (XD:
START) indrukken.
De display toont het resultaat in mg/L TSS (Total Suspended Solids).
Zwevende Vaste stoffen 24 / M384
NLMethod Reference Book1.0 45
NL
Chemische methode
Troebelheid / doorvallend licht
Aanhangsel
Verstoringen
Permanente verstoringen
Kleur stoort wanneer licht wordt geabsorbeerd bij 660 nm.
Uit te sluiten verstoringen
Luchtbellen interfereren en kunnen worden verwijderd door het spoelbakje een beetje
te draaien.
Validatie van de methodes
Aantoonbaarheidsgrens 10 mg/L
Bepaalbaarheidsgrens 30 mg/L
Einde meetbereik 750 mg/L
Gevoeligheid 550 mg/L / Abs
Betrouwbaarheidsgrenzen 4.24 mg/L
Standaardafwijking procedure 1.79 mg/L
Variatiecoefficient procedure 0.47 %
Afgeleid van
EN 872:2005
DE Methodenhandbuch 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
酸性 /指示剂
20
S:4.3
儀器的具體信息
測試可以在以下設備上執行。 此外還指出了所需的比色杯和光度計的吸收範圍。
仪器类型 比色皿 λ 测量范围
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
材料
所需材料(部分可選):
标题 包装单位 货号
Alka-M-Photometer 片剂 / 100 513210BT
Alka-M-Photometer 片剂 / 250 513211BT
应用列表
污水处理
饮用水处理
原水处理
备注
1. 术语碱度-mm-值、总碱度和酸容量 KS4.3 是相同的。
2. 准确地遵守 10 ml 的样本体积对分析结果的准确度至关重要。
屏幕显示: MD 100 /
MD 110 / MD 200
用于方法检测的条形码
方法名称
测量范围
化学方法
方法号
语言代码ISO 639-1 修订状态
KS4.3 T / 20
CN 方法手册 01/20
How to use
ENMethod Reference Book46
ZH
进行测定 KS4.3 片剂酸容量
选择设备中的方法。
对于这种方法,在以下设备上不能进行 ZERO 测量:XD 7000, XD 7500
10 ml
10 ml 样本填充 24 mm
色杯。
密封比色杯。
Sample
将样本比色杯放入测量轴
中。注意定位。
加入 ALKA-M-PHOTOME-
TER 片剂。
24 mm
用轻微的扭转压碎片剂。 密封比色杯。
开始测量
KS4.3 T / 20
DE Methodenhandbuch 01/20CN 方法手册 01/20
How to use
ENMethod Reference Book 47
ZH
悬浮固体 24 / M384
ZHMethod Reference Book1.048
ZH
悬浮固体 24 M384
10 - 750 mg/L TSS SuS
浊度 / 衰减辐射法
材料
所需材料(部分可選):
试剂 包装单位 货号
不需要试剂
取样
1. 取样后尽快测试水样。样本可以在 4 °C 下在塑料或玻璃瓶中储存长达 7 天。测量应
在与取样相同的温度下进行。测量和取样之间的温差可能会改变测量结果。
备注
1. 悬浮固体的光度测定基于重量分析法。在实验室中,过滤水样的过滤残余物的蒸馏通
常在 103 °C - 105 °C 的烘箱中进行,称出干燥的残余物。
2. 如果需要提高准确度,则应进行样本重量测定。这个结果可以用于用户调整同一样本
的光度计。
3. 该方法的估计检出限为 20 mg/L TSS。
悬浮固体 24 / M384
ZHMethod Reference Book1.0 49
ZH
进行测定 悬浮固体
选择设备中的方法。
在高速混合器中将 mL 水样均质化 分钟。
10 mL
Sample
10 mL 去离子水填充
24 mm 比色杯。
密封比色杯。 将样本比色杯放入测量轴
中。注意定位。
Zero
按下 ZERO 按钮。 从测量轴上取下比色杯。
24 mm
24 mm
10 mL
彻底混匀均匀的水样。 用水样预清洗比色杯。 10 mL 准备好的样本填充
24 mm 比色杯。
悬浮固体 24 / M384
ZHMethod Reference Book1.050
ZH
Sample
Test
样本比色杯放入测量轴
中。注意定位。
按下 TEST (XD: START)
按钮。
结果在显示屏上显示为 mg / l TSS (总固体悬浮物)。
悬浮固体 24 / M384
ZHMethod Reference Book1.0 51
ZH
化学方法
浊度 / 衰减辐射法
附錄
干扰说明
持续干扰
当光在 660 nm 处被吸收时,颜色会干扰。
可消除干扰
• 气泡干扰,可以通过轻微摇动比色杯来消除。
方法验证
检出限 10 mg/L
测定下限 30 mg/L
测量上限 750 mg/L
灵敏度 550 mg/L / Abs
置信范围 4.24 mg/L
标准偏差 1.79 mg/L
变异系数 0.47 %
源于
EN 872:2005
Tintometer GmbH
Lovibond® Water Testing
Schleefstraße 8-12
44287 Dortmund
Tel.: +49 (0)231/94510-0
www.lovibond.com
Germany
The Tintometer Limited
Lovibond House
Sun Rise Way
Amesbury, SP4 7GR
Tel.: +44 (0)1980 664800
Fax: +44 (0)1980 625412
www.lovibond.com
UK
Tintometer China
9F, SOHO II C.
No.9 Guanghualu,
Chaoyang District,
Beijing, 100020
Customer Care China Tel.: 4009021628
Tel.: +86 10 85251111 Ext. 330
Fax: +86 10 85251001
www.lovibond.com
China
Tintometer South East Asia
Unit B-3-12, BBT One Boulevard,
Lebuh Nilam 2, Bandar Bukit Tinggi,
Klang, 41200, Selangor D.E
Tel.: +60 (0)3 3325 2285/6
Fax: +60 (0)3 3325 2287
www.lovibond.com
Malaysia
Tintometer Brazil
Caixa Postal: 271
CEP: 13201-970
Jundiaí – SP
Tel.: +55 (11) 3230-6410
www.lovibond.com.br
Brazil
Tintometer Inc.
6456 Parkland Drive
Sarasota, FL 34243
Tel: 941.756.6410
Fax: 941.727.9654
www.lovibond.us
USA
Tintometer India Pvt. Ltd.
Door No: 7-2-C-14, 2nd, 3rd & 4th Floor
Sanathnagar Industrial Estate,
Hyderabad, 500018
Telangana
Tel: +91 (0) 40 23883300
Toll Free: 1 800 599 3891/ 3892
www.lovibondwater.in
India
Tintometer Spain
Postbox: 24047
08080 Barcelona
Tel.: +34 661 606 770
www.lovibond.com
Spain
Technical changes without notice
Printed in Germany 12/23
No.: 00386459
Lovibond® and Tintometer® are Trademarks of
the Tintometer Group of Companies
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55

Lovibond 00386459 - MD1x0/200 - Suspended solids 24 Manual de usuario

Tipo
Manual de usuario